JDS Uniphase Stud Sensor RM3750 User Manual

RM3 SERIES BACKREFLECTION METER  
USER’S MANUAL  
RM3 SERIES  
BACKREFLECTION METER  
User’s Manual  
10112341 Rev 002  
Page 1 of 39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RM3 SERIES BACKREFLECTION METER  
USER’S MANUAL  
Front Panel....................................................................................................................22  
Rear Panel ....................................................................................................................23  
Termination Techniques................................................................................................23  
Termination Techniques for Single-mode Fiber (except 980 nm) ......................23  
Termination Techniques for Single-mode (980 nm) and Multimode Fiber .........24  
Backreflection Measurements.......................................................................................25  
Setting Up the Meter for Backreflection Measurements.....................................25  
Measuring Backreflection ..................................................................................26  
Backreflection Accuracy and Range..................................................................27  
Loss and Power Measurements....................................................................................27  
Setting Up the Meter for Loss and Power Measurements .................................27  
Measuring Power...............................................................................................28  
Measuring Relative Power (Insertion Loss)........................................................29  
Power Accuracy.................................................................................................30  
Data Logging Using the Foot Pedal ..............................................................................30  
Messages and Symbols ................................................................................................30  
Calibrating the Meter.....................................................................................................31  
Backreflection Calibration Adjustment ...............................................................32  
Power Calibration Adjustment............................................................................32  
Maintaining the Meter....................................................................................................33  
Cleaning the Unit...............................................................................................33  
Cleaning the Connector Ends............................................................................33  
Recharging the Battery......................................................................................34  
Replacement of Internal Fuses..........................................................................34  
Troubleshooting.................................................................................................34  
Programming Guide................................................................................................................37  
RS232 Serial Interface..................................................................................................37  
RS232 Pin Assignment......................................................................................37  
RS232 Interface Specifications..........................................................................37  
RS232-GPIB Converter.................................................................................................39  
Left Dip Switch...................................................................................................39  
Right Dip Switch ................................................................................................39  
Programming Examples................................................................................................40  
RS232 Interface Program Example ...................................................................40  
Customized Features and Test Data......................................................................................41  
For sales and service information,  
contact JDS Uniphase or your local representative.  
JDS Uniphase Corporation  
570 West Hunt Club Road  
Nepean, Ontario, Canada  
K2G 5W8  
Phone: 613 727-1303  
Fax: 613 727-8284  
10112341 Rev 002  
Page 3 of 39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RM3 SERIES BACKREFLECTION METER  
USER’S MANUAL  
Safety Information, Instructions, and Symbols  
Safety Information  
Classification  
The RM Series Backreflection Meter consists of an exposed metal chassis that is connected  
directly to earth via a power cord and, therefore, is classified as a Class 1 instrument.  
Disconnecting from Line Power  
Some of the circuits are powered whenever the unit is connected to the AC power source (line  
power). To ensure that the unit is not connected to the line power, disconnect the power cord  
from either the power inlet on the unit’s rear panel or from the AC line-power source  
(receptacle). The power cord must always be accessible from one of these points. If the unit is  
installed in a cabinet, the operator must be able to disconnect the unit from the line power by  
the system’s line-power switch.  
Line Power Requirements  
The RM meter operates from any single-phase AC power source that supplies between 100  
and 240 V at a frequency range of 50 to 60 Hz. The maximum power consumption of the RM  
meter is 25 VA. The unit can also be operated using its internal 6V DC rechargeable battery.  
Fuse Type  
The fuse type used by the unit is (5 x 20) mm, 1 A/250 V (fast).  
Laser Specifications  
Laser specifications are provided in Table 1.  
Table 1: Specifications  
Parameter  
Wavelength  
Single-mode (SM)  
Multimode (MM)  
980*, 1310, 1480, 1550, 1625,  
or 1650 ± 10 nm  
850, or 1310 ± 20 nm  
Class  
1
1
Fiber Type  
9/125 µm SM(*5/125 µm SM) 50/125 µm MM 62.5/125 µm  
MM  
Maximum Accessible  
Emission Level  
-15 dBm  
-23 dBm  
-25 dBm  
Maximum Output Power  
Numerical Aperture  
-4 dBm  
-4 dBm  
0.20  
-4 dBm  
0.28  
0.13 (*0.14)  
0.20 (*0.21)  
Effective Numerical  
Aperture  
0.27  
0.34  
10112341 Rev 002  
Page 4 of 39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RM3 SERIES BACKREFLECTION METER  
USER’S MANUAL  
Under the laser classification of the US Food and Drug Administration (FDA) Center for  
Devices and Radiological Health (CDRH), the laser contained in the unit is a Class 1 laser.  
CLASS 1  
LASER PRODUCT  
10112341 Rev 002  
Page 5 of 39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RM3 SERIES BACKREFLECTION METER  
USER’S MANUAL  
Safety Instructions  
The following safety instructions must be observed whenever the unit is operated, serviced, or  
repaired. Failure to comply with any of these instructions or with any precaution or warning  
contained in the user’s manual is in direct violation of the standards of design, manufacture,  
and intended use of the unit. JDS Uniphase assumes no liability for the customer’s failure to  
comply with any of these safety requirements.  
Before Initializing and Operating the Unit  
Inspect the unit for any signs of damage, and read the user’s manual thoroughly.  
Install the unit as specified in the Getting Started section.  
Ensure that the unit and any devices or cords connected to it are properly grounded.  
Operating the Unit  
Warning  
To avoid the risk of injury or death, always observe the following precautions  
before initializing the unit:  
To prevent hazardous radiation exposure, use only the controls or  
adjustments and perform only procedures specified for the laser source.  
If using a voltage-reducing autotransformer to power the unit, ensure that  
the common terminal connects to the earthed pole of the power source.  
Use only the type of power cord supplied with the unit.  
Connect the power cord only to a power outlet equipped with a protective  
earth contact. Never connect to an extension cord that is not equipped with  
this feature.  
Willfully interrupting the protective earth connection is prohibited.  
Never look into the end of an optical cable connected to an optical output  
device that is operating. Laser radiation is invisible, and direct exposure can  
severely injure the human eye. For more information, see the user’s manual  
of the laser source in use.  
Do not use the unit outdoors.  
To prevent potential fire or shock hazard, do not expose the unit to any  
source of excessive moisture.  
Do not operate the unit when its covers or panels have been removed.  
Do not interrupt the protective earth grounding. Any such action can lead to  
a potential shock hazard that can result in serious personal injury.  
Do not operate the unit if an interruption to the protective grounding is  
suspected. In this case, ensure that the unit remains inoperative.  
Use only the type of fuse specified by JDSU as appropriate for this unit. Do  
10112341 Rev 002  
Page 6 of 39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RM3 SERIES BACKREFLECTION METER  
USER’S MANUAL  
not use repaired fuses, and avoid any situations that can short-circuit the  
fuse.  
Unless absolutely necessary, do not attempt to adjust or perform any  
maintenance or repair procedure when the unit is opened and connected to  
a power source.  
Repairs are to be carried out only by a JDSU qualified technician.  
Do not attempt any adjustment, maintenance, or repair procedure to the  
unit’s internal mechanism if immediate first aid is not accessible.  
Disconnect the power cord from the unit before adding or removing any  
components.  
Operating the unit in the presence of flammable gases or fumes is  
extremely hazardous.  
Do not perform any operating or maintenance procedure that is not  
described in the user’s manual.  
Some of the unit’s capacitors can be charged even when the unit is not  
connected to the power source.  
If the equipment is used in a manner not specified by the manufacturer, the  
protection provided by the equipment may be impaired.  
10112341 Rev 002  
Page 7 of 39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RM3 SERIES BACKREFLECTION METER  
USER’S MANUAL  
Safety Symbols  
The following symbols and messages can be marked on the unit (Table 2). Observe all safety  
instructions that are associated with a symbol.  
Table 2: Safety Symbols  
Symbol  
Description  
Laser safety. See the user’s manual for instructions on handling and operating  
the unit safely.  
See the user’s manual for instructions on handling and operating the unit  
safely.  
Electrostatic discharge (ESD). See the user’s manual for instructions on  
handling and operating the unit safely.  
Frame or chassis terminal for electrical grounding within the unit.  
Protective conductor terminal for electrical grounding to the earth.  
The procedure can result in serious injury or loss of life if not carried out  
in proper compliance with all safety instructions. Ensure that all  
conditions necessary for safe handling and operation are met before  
proceeding.  
WARNING  
CAUTION  
The procedure can result in serious damage to or destruction of the unit if not  
carried out in compliance with all instructions for proper use. Ensure that all  
conditions necessary for safe handling and operation are met before  
proceeding.  
10112341 Rev 002  
Page 8 of 39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RM3 SERIES BACKREFLECTION METER  
USER’S MANUAL  
Compliance  
CE Compliance  
The unit has been designed and tested to comply with directive 73/23/EEC and its subsequent  
amendments by the European Community (EC or CE). The directive relates to electrical  
equipment designed for use within certain voltage limits. It ensures that electrical equipment is  
constructed with good engineering practice in safety matters.  
The unit has been designed and tested to comply with directive 89/336/EEC and its  
subsequent amendments. The directive relates to electromagnetic compatibility. It demands  
that electromagnetic disturbance does not exceed a prescribed level; that the equipment be  
immune to a prescribed level of ambient level of interference; that the equipment be protected  
against electrostatic discharges; and that the equipment be immune to all electrical shock  
wave disturbances. As of 1997, measures have been added to test for fire hazard, electric  
shock hazard, and also external exposure to other forms of energy.  
The requirements specified by directive 89/336/EEC are as follows. CE compliance requires  
that the manufacturer or its authorized representative established within the Community affix  
the CE conformity mark to the apparatus or else to the packaging, instructions for use, or  
guarantee certificate. The CE conformity mark shall consist of the letters CE as specified and  
the figures of the year in which the mark was affixed. This mark should, where appropriate, be  
accompanied by the distinctive letters used by the notified body issuing the CE type-  
examination certificate. Where the apparatus is the subject of other Directives providing for the  
CE conformity mark, the affixing of the CE mark shall also indicate conformity with the relevant  
requirements of those other Directives.  
FDA-CDRH Compliance  
Under the US Food and Drug Administration (FDA) Center for Devices and Radiological Health  
(CDRH), the unit complies with the Code of Federal Regulations (CFR), Title 21, Subchapter J,  
1050.pdf for more information.  
CSA / IEC Compliance  
The unit complies with certain standards of the Canadian Standards Association (CSA) and  
the International Electrotechnical Commission (IEC).  
The unit is in Installation Category (Overvoltage Category) II under IEC 664. IEC 664 relates to  
impulse voltage levels and insulation coordination. The particular category is defined as: local  
level, appliances, portable equipment, etc, with smaller transient overvoltages than Installation  
Category (Overvoltage Category) III.  
The unit is in the Pollution Degree 2 category under IEC 1010-1 and CAN/CSA-C22.2 No.  
1010.1. The IEC standard on Safety Requirements for Electrical Equipment for Measurement,  
Control, and Laboratory Use relates to insulation coordination. The CSA standard is on Safety  
Requirements for Electrical Equipment for Measurement Control, and Laboratory Use, Part I:  
General Requirements. The Pollution Degree 2 category is defined as follows: “Normally only  
10112341 Rev 002  
Page 9 of 39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RM3 SERIES BACKREFLECTION METER  
USER’S MANUAL  
non-conductive pollution occurs. Occasionally, however, a temporary conductivity caused by  
condensation must be expected.”  
10112341 Rev 002  
Page 10 of 39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RM3 SERIES BACKREFLECTION METER  
USER’S MANUAL  
General Information and Specifications  
General Information  
This user’s manual for the RM3 Series Backreflection Meter contains complete operating  
instructions. The RM meter can be ordered as a single-mode or multimode model. Software is  
bundled with the meter; for information on using the software, see the Multiple Connector Test  
System Software User’s Manual (document SD000318). The inspection report and a  
description of any customer-requested information are found in Customized Features and  
Test Data section.  
The RM meter (Figure 1) is a portable, direct-display instrument that measures backreflection,  
loss, and power of single-mode and multimode fiberoptic connectors, components, and  
systems. The RM meter is supplied with one or two internal laser sources for operation at 850,  
980,1310, 1480, 1550, 1625, or 1650 nm (not all combinations are available).  
Figure 1: RM Meter  
An important feature of the RM meter is the FC/APC ultra-low backreflection connector at its  
output port (SC/APC connector also offered). By switching hybrid jumpers connected to the  
FC/APC connector, the user can select the connector type required for a measurement quickly  
and easily, without limiting the backreflection range of the meter. The RM meter is supplied  
with a hybrid measurement jumper to be used exclusively for measurements and a hybrid  
calibrated jumper to be used exclusively for calibrations.  
The RM meter can be operated using an internal 6 V DC rechargeable battery.  
The RM meter includes a convenient foot pedal, enabling the user to trigger the meter to send  
measurement data to a computer or printer.  
The RM meter operates over the RS232 serial interface. Wherever the IEEE 488 GPIB parallel  
interface is mentioned, it is achieved via a RS232-to-GPIB converter that can be purchased as  
an accessory.  
10112341 Rev 002  
Page 11 of 39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RM3 SERIES BACKREFLECTION METER  
USER’S MANUAL  
Backreflection Measurements  
Reflections in optical systems can come from a number of sources. Primary sources include  
the fiber (Rayleigh backscatter) and Fresnel reflections that occur at the planar junction of two  
materials having different refractive indices, for example, connector and fiber endfaces,  
splices, bulk optic interfaces, and detector surfaces.  
Typically only Fresnel reflections are significant because transmitters are relatively insensitive  
to distributed reflections such as Rayleigh backscatter. Backreflection caused by Rayleigh  
backscatter varies with the length and type of fiber, and is only significant when measuring  
components with backreflections below -40 dB or with very long pigtails (multimode). However,  
Rayleigh backscatter can be a large contributor to backreflection in installed systems or when  
using long fiber.  
The internal switch and coupler of the RM meter enable the meter to measure the internal light  
source signal (Pin), the signal offset with no light (Pdark), and the total signal level from internal  
and external backreflections (Pbr).  
The RM meter first calculates the total backreflection from internal and external sources (BRtot),  
using the following equation, where CAL is the factory-set calibration factor of the meter:  
BRtot = 10 log (Pbr - Pdark)/(Pin - Pdark) - CAL  
[dB]  
The backreflection from external sources (BR) is then calculated using the following equation,  
where BR0 is the stored value of the total backreflection up to the device under test (DUT), and  
User CAL is the user-set calibration factor:  
BR  
BR = 10 log (10BR  
- 10  
) + User CAL  
[dB]  
tot/10  
0/10  
The RM meter then displays the value of BR.  
The BR measurement takes approximately one second to complete. Before this is done, the  
Pin and Pdark measurements are completed in approximately four seconds, during which time  
the backreflection display is locked and the light from the output port is blocked. The values  
are then updated every minute.  
Loss and Power Measurements  
The RM meter is equipped with a front-panel InGaAs detector for relative power (loss) and  
absolute power measurements. (Absolute power is referred to as power in this manual; relative  
power is always referred to as relative power.) Because the meter is capable of storing the  
dark signal from the detector, high-accuracy power measurements as low as -80 dBm for the  
single mode model (and –60 dBm for the multimode model) can be obtained.  
When making power measurements, the RM meter first measures two signals: the total signal  
level (Itot) and the dark signal from the detector (ID). The RM meter then calculates the power  
measurement using the following equation, where CAL is the factory-set calibration factor and  
User CAL is the user-set calibration factor:  
P = 10 log (Itot - ID) + CAL + User CAL  
[dB]  
10112341 Rev 002  
Page 12 of 39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RM3 SERIES BACKREFLECTION METER  
USER’S MANUAL  
Output Port  
The output port of the RM meter is equipped with an ultra-low backreflection APC connector.  
To prevent damage to the output port connector, a measurement jumper must be used for all  
measurements, even for measuring a component with an APC connector.  
Internal Light Sources  
The RM meter can be equipped with two internal laser sources. These are thermoelectrically  
cooled for added stability and are modulated at 10 kHz. Fast changes in ambient temperature  
can cause a source to become unstable. If this occurs, the meter automatically switches to a  
measuring mode in which Pin and Pdark are measured more frequently than every minute and  
the message “Source Unstable” is displayed. The light source power is monitored, and the  
meter returns to the standard measurement mode when the source has stabilized.  
Hybrid Jumpers  
The RM meter is supplied with a hybrid measurement jumper and a hybrid calibrated jumper.  
The measurement jumper has an APC connector at the input end. The other end on the  
standard measurement jumper is PC, but by changing measurement jumpers this end is user-  
selected, to be compatible with the input connector of the DUT. The calibrated jumper has an  
APC connector at the input end and a PC connector at the output end.  
Use the measurement jumper for measurement purposes only; use the calibrated jumper for  
calibration purposes only. The calibrated jumper has a label attached to it identifying it as the  
calibrated jumper.  
Coherence Length  
Reflected light from multiple reflections can change the backreflection measured by the  
RM meter. This variation typically shows up as noise or drift in the signal. The internal light  
source is designed to have low coherence length (typically less than 10 cm). Thus,  
interference effects are typically seen only between very closely spaced components such as  
non-contacting connectors.  
Electrical Requirements  
The RM meter is supplied with a rechargeable battery. When fully charged, the battery  
provides up to eight hours of operation at room temperature. For laboratory use, use the  
RM meter with the power cord. The letter B is shown on the RM meter front panel display when  
the battery power is low.  
Key Features  
Backreflection measurements to -75 dB for single mode model (-40 dB for multi mode  
model)  
Power measurements to -80 dBm for single mode model (-60 dBm for multi mode model)  
Convenient foot pedal for data logging  
MCTS application software  
Built-in light sources at 850, 980, 1310, 1480, 1550, 1625, or 1650 nm  
Compensation for extraneous backreflection for accurate backreflection measurements  
10112341 Rev 002  
Page 13 of 39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RM3 SERIES BACKREFLECTION METER  
USER’S MANUAL  
User-calibration mode  
RS232 serial and IEEE 488 GPIB parallel interface  
Convenient data logging via serial port to a computer or serial printer  
Direct display of measured backreflection, power, or insertion loss  
Applications  
Connector backreflection and loss testing  
Component testing  
Installation verification  
Quality assurance (QA) acceptance testing  
Standard Accessories  
AC power cord (country specific)  
Laser sources specified when ordered  
FC/PC-FC/APC or SC/PC-SC/APC hybrid measurement jumper  
FC/PC-FC/APC or SC/PC-SC/APC hybrid calibrated jumper  
FC-type (or SC-type) detector adapter  
Detector cap  
Foot pedal  
MCTS software and user’s manual  
User’s manual  
Carrying case (for field use)  
Certificate of Compliance (NIST Traceability Report available on request)  
Optional Accessories  
Variety of detector adapters  
Measurement jumpers with various user-selected connectors  
RS232-to-GPIB converter  
10112341 Rev 002  
Page 14 of 39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RM3 SERIES BACKREFLECTION METER  
USER’S MANUAL  
Specifications  
The following optical specifications describe the warranted characteristics of the unit ( Table 3).  
Supplementary specifications describe the typical non-warranted performance of the unit  
(Table 4).  
Table 3: Optical Specifications  
Parameter  
Operating wavelength  
Fiber Type  
Specification  
Single-mode  
Multimode  
980, 1310, 1480, 1550, 850, 1310 nm  
1625, or 1650 ±10 nm  
±20 nm  
*
1060 multimode  
for the 980 nm)  
Detector type  
2 mm diameter InGaAs  
3 mm diameter InGaAs  
Range  
backreflection  
power  
0 to -75 dB  
0 to -80 dBm  
0 to –40 dB1  
3 to -60 dBm  
Absolute accuracy  
backreflection  
power  
±1.0 dB2,3  
±1.0 dB4  
±0.25 dBm typical at -10  
±0.25 dBm typical at -10  
dBm6  
dBm5,6  
Relative accuracy  
backreflection  
power  
±0.4 dB3  
±0.7 dB4  
±0.15 dB (>5 dB loss)6,7  
±0.05 dB (<5 dB loss),  
±0.15 dB (>5 dB loss)5  
Output connector type  
Remote interface  
Panel-mounted,  
low-backreflection APC  
RS232 or IEEE 488 (GPIB)  
1
Reduced backreflection accuracy in the last 10dB of range based on termination effectiveness. Depending on  
the measurement setup, measurements with lower levels are possible at reduced accuracy.  
User-calibration supported.  
2
3
For a typical application, add ±0.4 dB for readings between -60 and -67 dB. Add ±0.8 dB for readings  
between -67 and -72 dB. Add ±1.5 dB for readings between -72 and -75 dB.  
Following the user-calibration procedure at the recommended interval. For simple reflections, such as flat-end  
connectors.  
Add ±0.1 dB between -70 and -80 dBm.  
Immediately after performing a dark measurement.  
4
5
6
7
Add ±0.1 dB between 0 to 3 dBm and between –35 and –40 dBm.  
10112341 Rev 002  
Page 15 of 39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RM3 SERIES BACKREFLECTION METER  
USER’S MANUAL  
* Flexcor is a trademark of CorningR.  
10112341 Rev 002  
Page 16 of 39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RM3 SERIES BACKREFLECTION METER  
USER’S MANUAL  
Table 4: Other Specifications  
Electrical  
Input voltage  
Power consumption  
Physical  
100 to 240 V AC, 50 to 60 Hz  
25 VA maximum  
Display  
16-character LCD  
Dimensions (W x H x D) 26 x 11 x 26 cm  
Weight  
4 kg  
Environmental  
Operating temperature  
Storage temperature  
Humidity  
0 to 40 °C  
-40 to 70 °C  
maximum 95% RH from 0 to 40 °C  
transportable  
Equipment positioning  
10112341 Rev 002  
Page 17 of 39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RM3 SERIES BACKREFLECTION METER  
USER’S MANUAL  
Getting Started  
The RM3 Series Backreflection Meter consists of the meter, an AC power cord, two jumpers,  
and a foot pedal.  
Before Initializing and Operating the Unit  
Inspect the unit for any signs of damage.  
Read the user’s manual thoroughly, and become familiar with all safety symbols and  
instructions to ensure that the unit is operated and maintained safely.  
If the unit is used in a manner not specified by JDS Uniphase, the protection provided by  
the unit may be impaired.  
Initial Inspection  
Warning  
To avoid electrical shock, do not initialize or operate the unit if it bears any sign  
of damage to any portion of its exterior surface, such as the outer cover or  
panels.  
Check that the unit and contents are complete:  
1. Inspect the shipping container for any indication of excessive shock to the contents, and  
inspect the contents to ensure that the shipment is complete.  
2. Inspect the unit for structural damage that can have occurred during shipping.  
3. Connect the unit to the power source using the power cord provided.  
4. Set the power switch to ON to initialize the RM meter, and observe the power-up  
sequence. The model number of the meter is displayed. Internal measurements of P in and  
Pdark are made. The message “Stabilizing” is displayed as the light source stabilizes. The  
backreflection value (BR) of the front panel connector is measured and displayed.  
5. Set the power switch to OFF and disconnect the RM meter.  
6. Connect the foot pedal to the Log port at the back of the meter if you intend to send data  
from the meter to a computer or printer.  
7. Keep the packaging.  
Immediately inform JDS Uniphase and, if necessary, the carrier if the contents of the shipment  
are incomplete, if the unit or any of its components are damaged or defective, or if the unit  
does not pass the initial inspection.  
10112341 Rev 002  
Page 18 of 39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RM3 SERIES BACKREFLECTION METER  
USER’S MANUAL  
Operating Environment  
In order for the unit to meet the warranted specifications, the operating environment must meet  
the following conditions for altitude, temperature, humidity, and voltage.  
Altitude  
The unit can be operated at an altitude up to 2000m.  
Temperature  
The unit can be operated in the temperature range of 0 to 40 °C.  
Humidity  
The unit can be operated in environments with up to 95% humidity (0 to 40 °C). Do not expose  
it to any environmental conditions or changes to environmental conditions that can cause  
condensation to form inside the unit.  
Voltage  
The main supply voltage fluctuations are not exceeding ±10% of the nominal voltage.  
Warning  
Do not use the unit outdoors.  
To prevent potential fire or shock hazard, do not expose the unit to any  
source of excessive moisture.  
Storing and Shipping  
To maintain optimum operating reliability, do not store the unit in locations where the  
temperature falls below -40 °C or rises above 70 °C. Avoid any environmental condition that  
can result in internal condensation. Ensure that these temperature and humidity requirements  
can also be met whenever the unit is shipped.  
Claims and Repackaging  
Immediately inform JDS Uniphase and, if necessary, the carrier, if  
The contents of the shipment are incomplete  
The unit or any of its components are damaged or defective  
The unit does not pass the initial inspection  
In the event of carrier responsibility, JDS Uniphase will allow for the repair or replacement of  
the unit while a claim against the carrier is being processed.  
Returning Shipments to JDS Uniphase  
JDS Uniphase only accepts returns for which an approved Return Material Authorization  
(RMA) has been issued by JDS Uniphase sales personnel. This number must be obtained prior  
to shipping any material to JDS Uniphase. The owner’s name and address, the model number  
and full serial number of the unit, the RMA number, the original JDSU sales order number, the  
10112341 Rev 002  
Page 19 of 39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RM3 SERIES BACKREFLECTION METER  
USER’S MANUAL  
invoice number, and an itemized statement of claimed defects must be included with the return  
material.  
Ship return material in the original shipping container and packing material. If these are not  
available, packaging guidelines are as follows:  
1. Cover the front panel with foam to prevent damage.  
2. Wrap the unit in anti-static packaging. Use anti-static connector covers.  
3. Pack the unit in a reliable shipping container.  
4. Use enough shock-absorbing material (10 to 15 cm) to cushion the unit and prevent it from  
moving inside the container. Pink poly anti-static foam is the best material.  
5. Seal the shipping container securely.  
6. Clearly mark FRAGILE on its surface.  
7. Always provide the model and serial number of the unit and, if necessary, the RMA number  
on any accompanying documentation.  
8. In order to ship the unit please contact the JDS RMA department at 613-727-1303.  
Cleaning Connectors  
Caution  
Connecting damaged or dirty fibers to the unit can damage the connectors  
on the unit.  
Never force an optical connector. Some connectors have a ceramic ferrule  
that can easily be broken.  
Optical cable ends need to be cleaned before using them with the unit.  
The following items are required for cleaning:  
Filtered compressed air or dusting gas (for example, Tech Spray Envi-Ro-Tech Duster  
1671 gas, available from http://www.techspray.com)  
Lint-free pipe cleaners (for example, from 3M1) or lint-free swab  
Lint-free towels (for example, 10 x 10 cm or 4 x 4 in HydroSorb III wipers, available from  
http://www.focenter.com)  
Optical grade isopropyl alcohol or optical grade 200° ethanol (do not use rubbing alcohol,  
which contains 30% water)  
To clean the connectors:  
1 3M is a trademark of 3M.  
10112341 Rev 002  
Page 20 of 39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RM3 SERIES BACKREFLECTION METER  
USER’S MANUAL  
1. Blow the sleeve with filtered compressed air (Figure 2).  
sleeve  
ferrule  
Figure 2: Connector Cleaning (connector type can vary)  
2. Apply optical grade isopropyl alcohol or optical grade ethanol (do not use rubbing alcohol)  
to a small area of a lint-free towel and rub the end of the ferrule over the wet area.  
3. Wipe the ferrule on a dry area of the lint-free towel.  
4. Using the dusting gas or compressed air, blow the end of the ferrule.  
5. Apply the alcohol or ethanol to a lint-free pipe cleaner or swab and wipe off the remaining  
parts of the connector.  
6. With the other end of the pipe cleaner or swab, dry the areas cleaned.  
7. Using the dusting gas or compressed air, blow the areas cleaned.  
Powering Up the Meter  
To power up the meter:  
1. Connect the meter to an AC power source using the power cord provided.  
2. Set the power switch to ON to initialize the RM meter, and observe the power-up  
sequence: The model number of the meter is displayed. Internal measurements of P in and  
Pdark are made. The message “Stabilizing” is displayed as the light source stabilizes. The  
backreflection value (BR) of the front panel connector is measured and displayed.  
10112341 Rev 002  
Page 21 of 39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RM3 SERIES BACKREFLECTION METER  
USER’S MANUAL  
Operating and Maintenance Instructions  
Front Panel  
The front of the meter (model RM3750) is shown in Figure 3 and described inTable 5. Not all  
RM3 Series meters are exactly as shown.  
Figure 3: Front of the Meter  
Table 5: Front Panel Operating Keys and LEDs  
Key/LED  
ON / OFF  
Description  
Power on ON / OFF switch.  
BR  
Sets the meter to Backreflection measurement mode.  
STORE BR0  
Stores the BR0 value used in backreflection measurements and toggles  
the key lamp on and off.  
POWER  
STORE ID  
λ
Sets the meter to Power and Relative Power measurement modes.  
Stores the ID value used in dark measurements.  
Selects one of the two laser sources.  
10112341 Rev 002  
Page 22 of 39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RM3 SERIES BACKREFLECTION METER  
USER’S MANUAL  
Rear Panel  
The back of the meter (Figure 4) can vary with the model.  
-
25 VA max  
Fuse 1A/250V  
Figure 4: Back of the Meter  
The components on the rear panel are described in Table 6.  
Table 6: Rear Panel Components  
Component  
CAL  
Function  
Access to the User CAL value, for calibration purposes.  
Foot pedal jack for data logging.  
LOG  
RS232C  
RS232 serial interface port.  
Termination Techniques  
Termination is a technique used to block all backreflections beyond a certain point in a network  
of components. In all measurement testing, two terminations must be made: the first (for  
measuring BR0) is made before the DUT, for example, on the RM meter side; the second (for  
measuring BRtot) is made after the DUT. Both termination points need to be made as close as  
possible to the DUT to minimize the errors associated with Rayleigh backscattering in the fiber.  
The backreflection of all components and connections between these two points is included in  
the equation for calculating the backreflection value (BR) of the DUTs.  
Termination Techniques for Single-mode Fiber (except 980 nm)  
Termination for single-mode (except 980nm) is performed by mandrel wrapping.  
High-attenuation bends in the fiber (bends with a relatively small radius) remove all  
backreflection and can be made anywhere along the length of the cable. When measuring  
backreflection levels below -60 dB, low-attenuation bends (bends with a relatively large radius)  
10112341 Rev 002  
Page 23 of 39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RM3 SERIES BACKREFLECTION METER  
USER’S MANUAL  
need to be made just before the high-attenuation bends in order to offset the small amount of  
reflection caused by the high-attenuation bends.  
1. Using the size of rod specified in the Inspection Report (see the Customized Features  
and Test Data section), wind the cable around the rod until the reading displayed on the  
RM meter no longer changes (approximately six turns). The diameter of the rod must be  
suitable for the type of fiber connected to the output port jumper and for the wavelength in  
use. To minimize any memory of the bends in the cable jacket, do not pull on the cable  
while winding, and occasionally wind the cable in the opposite direction (figure 5a).  
Figure 5a: Terminating the Jumper for BR Measurement for Single-mode (except 980nm)  
2. The second termination is made after the DUT in order to obtain Brtot (figure 6a).  
Figure 6a: Measuring Backreflection for Single-mode (except 980 nm)  
Termination Techniques for Single-mode (980 nm) and Multimode Fiber  
Termination is performed by using index-matching gel.  
1. Apply index-matching gel directly to the fiber end (figure 5b).  
2. After measuring BR0, clean the connector end in accordance with the Cleaning  
Connectors section.  
10112341 Rev 002  
Page 24 of 39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RM3 SERIES BACKREFLECTION METER  
USER’S MANUAL  
Figure 5b: Terminating the Jumper for BR Measurement for Single-mode (980 nm only) and  
Multimode  
3. The second termination is made after the DUT in order to obtain Brtot (figure 6b).  
Figure 6b: Measuring Backreflection for Single-mode (980 nm only) and Multimode  
Backreflection Measurements  
Setting Up the Meter for Backreflection Measurements  
To prepare the meter for backreflection measurements:  
1. Ensure that the meter is powered off.  
2. Set the power switch to ON, and allow the meter to stabilize.  
3. Press the BR key to set the meter to Backreflection mode, which is displayed as BR on the  
meter.  
4. Clean the output port on the front of the meter and the APC connector (green boot) of the  
measurement jumper, and connect this end of the jumper to the output port. Ensure that  
you are using the measurement jumper and not the calibration jumper, which is labeled as  
such. The output connector of the measurement jumper is user-selected and must be  
compatible with the input connector of the DUT.  
For each wavelength at which measurements are to be made:  
5. Press the λ key to select the required wavelength.  
10112341 Rev 002  
Page 25 of 39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RM3 SERIES BACKREFLECTION METER  
USER’S MANUAL  
6. Terminate the measurement jumper just before the output connector, and hold the  
termination point steady (figure 5a for single-mode(except 980nm), and figure 5b for single-  
mode(980nm) and multimode).  
NOTE  
The termination technique is different depending on the type of the RM Meter used :  
single-mode (except 980 nm), single-mode (980nm) or multimode.  
Please refer to the Termination Techniques section above.  
7. Press the Store BR0 key. The key lamp lights to indicate that the background backreflection  
has been stored. The BR0 value is used in the calculation of the DUT backreflection at the  
selected wavelength.  
8. Remove the termination.  
Measuring Backreflection  
To measure backreflection and power from a unit after setting up the meter (see the Setting  
Up the Meter for Backreflection Measurements section):  
1. Clean the output connector of the measurement jumper and the input connector of the  
DUT, and mate the two.  
2. Press the λ key to select the wavelength at which the measurement is to be made.  
3. Terminate the measurement jumper immediately after the DUT, and hold the termination  
point steady. The meter displays the backreflection between the two termination points that  
is caused by the fiber, all connections, and the DUT. This area is shown within the dashed  
line in Figure 6a for single-mode (except 980nm), and in Figure 6b for single-mode  
(980nm) and multimode.  
4. If necessary, terminate the fiber for BR0 again just before the DUT to eliminate the  
backreflection caused by the fiber and all connections before the DUT.  
5. Press the Store BR0 key. The key lamp lights to indicate that the function is on.  
6. If the backreflection setup has been done previously (BR 0 light is on), press the BR0 button  
to delete the currently stored BR0 value. The BR0 light turns off.  
7. Press λ and repeat step 6 to delete the stored BR0 value for another wavelength.  
8. Release the termination. The meter now displays the backreflection of the DUT only.  
9. If a measurement is to be made at a second wavelength, repeat steps 1 through 8.  
The small amount of reflections that BR0 represents can be polarization sensitive, and multiple  
reflections can cause interference effects that can make the reflections very sensitive to  
10112341 Rev 002  
Page 26 of 39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RM3 SERIES BACKREFLECTION METER  
USER’S MANUAL  
temperature. To ensure accurate backreflection measurements below -65 dB (-25 dB for  
multimode):  
1. Perform the BR0 measurement at least once per shift.  
The measured backreflection is affected by losses that can occur between the meter and the  
DUT. To compensate for these losses, measure the total amount of loss between the output of  
the measurement hybrid jumper and the input of the DUT (see the Loss and Power  
Measurements section). Double this value and add it to the backreflection value displayed on  
the meter. In the following example, the total loss between the meter and the DUT is 2.0 dB,  
and the displayed backreflection is -29 dB:  
BR = -29 dB + (2 x 2.0 dB) = -25 dB  
Backreflection Accuracy and Range  
The absolute accuracy of backreflection measurements made with the meter is dependent on  
the level of backreflection to be measured.  
The backreflection measurement range is restricted by the BR0 level. The meter can measure  
backreflection levels 15 dB below BR0 to a maximum of -75 dB (-40dB for multimode). For  
example, if BR0 is -40 dB, the minimum backreflection of the DUT that can be measured is -55  
dB.  
An asterisk (*) is displayed near the range limit (last 5 dB of range if using BR0, below -60 dB  
otherwise) to indicate that the setup and measurement procedures described in the previous  
two sections must be followed carefully to ensure accurate results.  
Absolute accuracy specifications are dependent upon the accuracy of the RM meter  
calibration, so a calibration check needs to be performed periodically (see the Calibrating the  
Meter section).  
Loss and Power Measurements  
Setting Up the Meter for Loss and Power Measurements  
To prepare the meter for loss and power measurements:  
1. Ensure that the RM meter is powered off.  
2. Set the power switch to ON, and allow the RM meter to stabilize.  
3. Clean the output port on the front of the meter and the FC/APC connector (green boot) of  
the measurement jumper, and connect this end of the jumper to the output port. Ensure  
that you are using the measurement jumper and not the calibration jumper, which is  
labeled as such.  
For each wavelength at which measurements are to be made:  
4. Press the λ key to select the required wavelength.  
10112341 Rev 002  
Page 27 of 39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RM3 SERIES BACKREFLECTION METER  
USER’S MANUAL  
5. Attach the appropriate detector adapter to the detector on the front of the meter.  
6. Connect the output connector of the measurement jumper to the detector adapter. This  
end of the measurement jumper is user-selected and must be compatible with the input  
connector of the DUT.  
7. Press the Power key to set the meter to Power mode. When in Power mode, the meter  
displays P at the beginning of the reading, and no “rel” is displayed, for example, P = -9  
dBm.  
8. Press and hold the Power key until the display on the meter clears.  
9. Release the Power key. The display reads 0.00 dB, and the loss of the measurement  
jumper is subtracted from the DUT loss or power measurements to be made (Figure 7).  
Figure 7: Loss and Power Measurements  
For loss and power measurements below -50 dBm:  
10. Ensure that the detector is covered, for example, that the detector cap is on or the output  
connector is connected to the detector. This dark measurement current is subtracted from  
future loss and power measurements.  
11. Press the Store ID key. The key lamp lights to indicate that the user-stored value of ID (the  
dark signal from extraneous sources) is used to calculate the DUT loss or power. If the  
detector is not sufficiently covered, the message “Too Much Light” is displayed. If the ID  
value is not stored before a measurement is made, the last value stored is used.  
Measuring Power  
In Power mode, absolute power measurements are displayed in dBm. To make a power  
measurement of a DUT after performing the setup steps in the Setting Up the Meter for Loss  
and Power Measurements section:  
1. Press the Power key to set the meter to Power mode.  
2. Press the λ key to select the wavelength at which the measurement is to be done.  
3. Clean the output connector of the measurement jumper and the input connector of the  
DUT, and connect the two connectors.  
10112341 Rev 002  
Page 28 of 39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RM3 SERIES BACKREFLECTION METER  
USER’S MANUAL  
4. Connect the output connector of the DUT to the detector on the front of the meter. The  
detector must be equipped with the appropriate detector adapter. The RM meter displays  
the DUT absolute power output (Figure 8).  
5. If a measurement is to be made at a second wavelength, repeat steps 2 to 4.  
Figure 8: Measuring Absolute Power  
Measuring Relative Power (Insertion Loss)  
The value for relative power is the negative value of the insertion loss, for example -31 dB  
instead of 31 dB. Relative power is displayed by the meter in dB.  
To make a relative power measurement of a DUT after performing the setup steps in the  
Setting Up the Meter for Loss and Power Measurements section:  
1. Press the Power key to set the meter to Relative Power mode. In Relative Power mode, the  
meter displays P at the beginning of the reading and rel at the end, for example, P = -0.15  
dB rel.  
2. Press the λ key to select the wavelength at which the measurement is to be made.  
3. Clean the output connector of the measurement jumper and the input connector of the  
DUT, and connect the two connectors. Connect the output connector of the DUT to the  
detector on the front of the meter. The meter displays the loss due to the DUT (Figure 9).  
4. Press the λ key to measure and display the loss at another wavelength.  
10112341 Rev 002  
Page 29 of 39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RM3 SERIES BACKREFLECTION METER  
USER’S MANUAL  
Figure 9: Measuring Relative Power  
Power Accuracy  
The absolute accuracy of power measurements made with the RM meter is dependent on the  
power level to be measured. The RM meter does not measure power levels below -50 dBm  
unless the dark signal from extraneous sources (ID) is stored.  
To ensure accurate power measurements below -65 dBm, store ID frequently.  
The value of ID can change with temperature fluctuations.  
Data Logging Using the Foot Pedal  
The RM meter is supplied with a foot pedal for convenient data logging to either a computer or  
printer. Data logging operates at a baud rate of 2400.  
To use the foot pedal:  
1. Ensure the computer or printer is connected to the RS232C or IEEE 488 (GPIB) port at the  
back of the meter.  
2. If the foot pedal is not already operational, connect the foot pedal plug to the Log jack on  
the rear panel of the meter.  
3. Press the foot pedal. The message “Log” appears on the right-hand side of the front-panel  
display, and the most recent measurements made are sent to the computer or printer  
through the RS232 serial or IEEE 488 (GPIB) port.  
Messages and Symbols  
The messages and symbols displayed by the RM meter are described in Table 7.  
Table 7: RM Meter Display Messages and Symbols  
Display  
Description  
RMXXXXB VERY.Y  
Displayed momentarily during the power-up sequence and indicates  
10112341 Rev 002  
Page 30 of 39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RM3 SERIES BACKREFLECTION METER  
USER’S MANUAL  
Display  
Description  
the model number (XXXX) and the software version (Y.Y).  
STABILIZING  
Displayed momentarily during the power-up sequence as the initial  
internal reference measurements are made.  
BR=-XX.XdB  
P=-XX.XXdBm  
P=-XX.XXdB rel  
*
Backreflection mode.  
Power mode.  
Relative Power mode.  
Displayed when the BR value is within the last 5 dB of range if using  
BR0, or from -60 to -75 dB if not using BR0.  
B
Low battery.  
<
P or BR is lower than the minimum value.  
P is greater than the maximum value.  
>
BRo STORED=-XX.X  
Displayed momentarily when the BR value (-XX.XdB) is being stored  
as BR0.  
ID STORED  
Displayed momentarily when measuring I D.  
TOO MUCH LIGHT  
INPUT LOW  
Displayed momentarily if the ID value is too high.  
Displayed momentarily when the laser power is too low.  
Displayed momentarily when the laser power is too high.  
Displayed momentarily when the laser power is unstable.  
INPUT HIGH  
SOURCE UNSTABLE  
1.3 BR CAL= X.X  
Indicates the backreflection user calibration value at 1.3 µm (1.3 µm  
is taken as an example).  
1.5 BR CAL= X.X  
1.3 PW CAL= X.XX  
1.5 PW CAL= X.XX  
Indicates the backreflection user calibration value at 1.5 µm.  
Indicates the power user calibration value at 1.3 µm.  
Indicates the power user calibration value at 1.5 µm.  
Calibrating the Meter  
Caution  
Devices with malfunctioning lasers must be returned to the manufacturer for  
calibration.  
Annual calibration of the meter by JDS Uniphase is recommended. This section outlines how  
to do periodic adjustments to ensure the accuracy of backreflection, loss, and power  
measurements. Use only the calibrated jumper for all calibration procedures.  
10112341 Rev 002  
Page 31 of 39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RM3 SERIES BACKREFLECTION METER  
USER’S MANUAL  
Backreflection Calibration Adjustment  
The RM meter has a factory-set backreflection calibration value (User CAL) that compensates  
for the typical loss resulting from a connection to the front-panel FC/APC connector. This value  
appears in the Inspection Report (see theCustomized Features and Test Data section).  
To perform a backreflection calibration adjustment:  
1. Press the BR key to set the RM meter to Backreflection mode, which is displayed as BR on  
the meter.  
2. Clean and connect the APC connector (green boot) of the calibrated jumper to the output  
port of the RM meter.  
3. Press the λ key to select the wavelength at which the calibration check is to be performed.  
Choose either wavelength, as you will check both.  
4. Clean the output connector of the calibrated jumper. The backreflection value is displayed.  
The value displayed on the RM meter must match the reference value that appears in the  
inspection report and on the label of the calibrated jumper. If these values do not match,  
ensure that the FC/APC connector is properly connected to the RM meter, and clean all the  
connectors. If the values still do not match to within ±0.3 dB, adjust the value of User CAL. The  
CAL key is located on the rear panel of the RM meter.  
To adjust the value of User CAL:  
1. Use a screwdriver or an appropriately narrow object to press the CAL key on the back of  
the RM meter. The last User CAL value stored is displayed on the RM meter.  
2. Press and hold the Store BR0 key to scroll to the required value in 0.1 dB increments.  
3. Press the CAL key to store the User CAL value and return the RM meter to the  
backreflection measurement display.  
Perform calibration adjustment for the second wavelength:  
1. Repeat the calibration adjustment procedure for the second wavelength.  
Power Calibration Adjustment  
To perform a power calibration adjustment:  
1. Press the Power key to set the RM meter to Power mode. In Power mode, the meter  
displays P at the beginning of the reading, and the value displayed is not followed by “rel”.  
An example is P = -9 dBm.  
2. Clean and connect the FC/APC connector (green boot) of the calibrated jumper to the  
output port of the meter.  
10112341 Rev 002  
Page 32 of 39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RM3 SERIES BACKREFLECTION METER  
USER’S MANUAL  
3. Press the λ key to select the wavelength at which the calibration check is to be performed.  
Choose either wavelength, as you will check both.  
4. Measure the power from the output of the calibrated jumper, using a high-accuracy,  
calibrated power meter.  
5. Clean the output connector of the calibrated jumper, and connect it to the detector on the  
front panel of the RM meter. The value displayed on the RM meter must match the value  
measured by the power meter to within ±0.3 dBm.  
If the values do not match to within ±0.3 dBm, adjust the value of User CAL:  
1. Use a screwdriver or an appropriately narrow object to press CAL. The last User CAL value  
stored is displayed on the meter.  
2. Press and hold the Store ID key to scroll to the required value in 0.01 dBm increments.  
3. Press the CAL key to store the User CAL value, and return the meter to the power  
measurement display.  
4. Repeat the calibration adjustment procedure (steps 1 to 3) for the second wavelength.  
Maintaining the Meter  
Caution  
Devices with malfunctioning lasers must be returned to the manufacturer for  
repair.  
Cleaning the Unit  
1. Unplug the unit from the line power.  
2. Clean the enclosure with a damp cloth.  
3. DO NOT plug the unit back until it is completely dry.  
Cleaning the Connector Ends  
Clean all connector ends with a lint-free tissue and alcohol before every mating. See the  
Cleaning Connectors section.  
1. Loosen the retaining screws of the connector panel, and remove the panel carefully to  
access the internal connectors (Figure 10).  
10112341 Rev 002  
Page 33 of 39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RM3 SERIES BACKREFLECTION METER  
USER’S MANUAL  
Figure 10: Removing the Connector Panel  
2. Remove the connector from the mating sleeve in the panel.  
3. Clean the connector endfaces and mating sleeve in accordance with the Cleaning  
Connectors section.  
4. Reinstall the connector onto the panel.  
5. Reinstall the connector panel. To avoid damaging the input and output port fibers, make  
one or two large loops in the fibers when replacing the panel.  
Recharging the Battery  
The letter B is displayed on the RM meter when the battery power is low. To recharge the  
battery:  
1. Connect the meter to the AC source.  
The RM meter can be used while it is recharging.  
Replacement of Internal Fuses  
The following fuses can be replaced by JDSU qualified technicians only:  
1. Fuse F1 on battery protection circuit PCB, 250V, F2A.  
2. Fuse on battwery cord, 250V, F 1 6/10 A  
Troubleshooting  
If any problem described in this section persists, contact JDS Uniphase or your local  
representative.  
Every backreflection measurement is affected by the loss before the DUT and by reflections  
before and after the DUT.  
10112341 Rev 002  
Page 34 of 39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RM3 SERIES BACKREFLECTION METER  
USER’S MANUAL  
Loss Before the DUT  
When a DUT is connected to the RM meter, the loss from connectors before the DUT affects  
the backreflection reading. Because the light is going out from the RM meter and then  
returning, it goes through the connectors twice, so the effect of the loss is doubled. If the loss  
is close to 0.15 dB, its effect is compensated automatically by the meter. If the loss differs  
substantially from 0.15 dB, follow the measurement procedures described at the end of the  
Measuring Backreflection section.  
Reflections Before the DUT  
When a device is connected to the RM meter, the reflections from the connectors affect the  
backreflection reading. The front panel FC/PC connections must have backreflection levels  
lower than -45 dB for single mode (-25 dB for multi mode), and FC/APC connections must  
have backreflection levels lower than -60 dB for single mode (-40 dB for multi mode). To  
ensure that the RM meter automatically compensates for this backreflection, follow the setup  
procedure described in the Setting Up the Meter for Backreflection Measurements section,  
and ensure that the backreflections do not change from the time of the setup.  
Reflections After the DUT  
When making a backreflection measurement, the fiber after the DUT must be terminated to  
eliminate the reflections from the end of the fiber (see the Termination Techniques section).  
For greatest accuracy when making very low backreflection measurements, terminate near the  
DUT to eliminate reflections from the fiber itself.  
Front Panel Connectors  
Follow the maintenance procedure described in the Maintaining the Meter section, to ensure  
that the internal pigtail connectors are clean and properly connected to the front panel mating  
sleeve.  
Connector Loss and Backreflection  
Ensure that the loss and backreflection of all the connectors are low and stable, complying  
with the connector specifications.  
Internal Backreflection (for single mode model only, except 980 nm model)  
To confirm proper operation of the RM meter, terminate the internal cable while using the  
factory-set BR0 value (BR0 key lamp is off). The backreflection reading must be less than  
-65 dB.  
Long Cables (for single mode model only, except 980 nm model)  
Terminate the cables near the devices or connectors being tested in order to eliminate  
backreflections from the fiber.  
10112341 Rev 002  
Page 35 of 39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RM3 SERIES BACKREFLECTION METER  
USER’S MANUAL  
Laser Stability  
If the messages “Source Unstable” or “Input Low” are displayed, remove the front connector  
panel, clean the front panel connectors, connect the internal FC/APC connector to the detector  
adapter, and press the Power key to set the meter to Power mode. The reading must be  
steady and higher than -10 dBm. If not, perform the procedure recommended in the table that  
follows.  
Calibration  
If the backreflection or power calibration was not performed properly, set BR CAL to 0.3 dB  
and Power CAL to 0 dB to restore the factory-set calibration. These settings provide  
reasonably accurate backreflection and power measurements.  
Other  
Display  
Problem  
Solution  
Meter does not turn on  
Set the power switch to OFF, connect  
the RM meter to a power source, and  
wait a few minutes before turning the  
power switch on.  
STABILIZING  
*
Message is displayed for a  
long time  
Set the power switch to OFF, connect  
the RM meter to a power source, and  
then set the power switch to ON.  
The BR measurement is  
approaching the range limit  
of the RM meter  
Follow the setup and measurement  
techniques described in the Setting  
Up the Meter for Backreflection  
Measurements and Measuring  
Backreflection sections.  
B
Battery is running low  
Connect the RM meter to the a power  
supply using the cord provided.  
<
BR range is very limited  
Follow the setup and measurement  
techniques described in the Setting  
Up the Meter for Backreflection  
Measurements and Measuring  
Backreflection sections.  
TOO MUCH LIGHT  
INPUT LOW  
ID value is too high  
Attach the detector cap to the detector,  
and press the ID key.  
Laser power is too low  
Check the measurement setup for an  
external source that entered into the  
output.  
INPUT HIGH  
Laser power is too high  
Check the measurement setup for an  
external source that entered into the  
output.  
10112341 Rev 002  
Page 36 of 39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RM3 SERIES BACKREFLECTION METER  
USER’S MANUAL  
Programming Guide  
RS232 Serial Interface  
The RM meter is equipped with an RS232 serial interface. Through the RS232 connection, the  
user can download measured values of backreflection, loss, and power to a computer or  
printer. The RM meter can also use an IEEE 488 parallel interface through the external RS232-  
to-IEEE 488 converter (described in the RS232-GPIB Converter section).  
The RS232 serial interface is configured as data terminal equipment (DTE). The data protocol  
is set to 2400 baud, ASCII character code with eight bits per character, two stop bits, and no  
parity bit.  
RS232 Pin Assignment  
The RS232 pin assignment is shown in Figure 11.  
Figure 11: RS232 Pin Assignment  
RS232 Interface Specifications  
The RS232 interface specifications are listed in Table 8.  
Table 8: RS232 Interface Specifications  
Name  
Symbol  
TxD  
Pin Number  
Signal Direction  
Transmitted data  
Received data  
Request to send  
Clear to send  
2
3
8
7
4
6
5
RxD  
RTS  
CTS  
DCD  
DTR  
SG  
Data carrier detect  
Data terminal ready  
Signal ground  
To connect the RM meter to the serial port of a computer:  
1. Attach a straight-through RS232 cable to the computer and the RS232C port at the back of  
the meter.  
10112341 Rev 002  
Page 37 of 39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RM3 SERIES BACKREFLECTION METER  
USER’S MANUAL  
On power-up, the RTS and DTR signals on the interface connector go positive, indicating that  
the interface is ready to communicate. The software on the controlling computer must enable  
the RTS line. The interface controller waits for the CTS line to go positive before the  
measurements are transmitted. The DCD line must go positive in order for the controlling  
computer to receive commands from the meter.  
The meter responds to single character commands sent on the serial interface. The characters  
must be sent individually, without a message terminating sequence such as <CR> (carriage  
return) or <LF> (line feed). The meter acknowledges each character that is sent with a  
character or a measurement reading.  
The command G initiates the transmission of a measurement. The meter returns the most  
recent measurement. The transmission also includes the wavelength at which the  
measurement was made and the characters <CR> <LF>.  
The character commands that control the meter via the RS232 serial interface port are listed in  
Table 9. When a command is accepted, the letter Y is returned. When a command is not  
recognized or not accepted, the letter X is returned.  
Table 9: Character Commands  
Command  
Action  
B
A
R
S
C
D
0
Display backreflection  
Display absolute power  
Display relative power  
1
Store BR0  
1
Clear BR0  
2
Store ID  
Set relative power to zero3  
3
Set wavelength to the laser connected to the first LED  
on display (eg. Model RM3750, first laser=1310nm. See  
Figure 3)  
5
Set wavelength to the laser connected to the second  
LED on display (eg. Model RM3750, second  
laser=1550nm. See Figure 3)  
1 Valid in Backreflection mode only.  
2 Valid in Absolute Power and Relative Power modes only.  
3 Valid in Relative Power mode only.  
10112341 Rev 002  
Page 38 of 39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RM3 SERIES BACKREFLECTION METER  
USER’S MANUAL  
RS232-GPIB Converter  
An optional user-supplied RS232-to-IEEE 488 converter can be used with the RS232  
connector on the RM meter to access an IEEE 488 parallel interface. The converter, supplied  
with the appropriate RS232 cable, is attached to the RS232 connector on the rear panel.  
The converter is set up and tested as described in Table 10 and Table 11. To avoid conflict  
with other GPIB devices, the GPIB address can be changed.  
Left Dip Switch  
The settings for the left dip switch are shown in Table 10.  
Table 10: Left Dip Switch Settings  
Switch  
GPIB ADDR  
CONFIG  
Setting  
01000  
Meaning  
address = 2  
ON  
use large serial buffer, enable  
SRQ  
MODE  
OFF  
device mode  
Right Dip Switch  
The settings for the right dip switch are shown in Table 11.  
Table 11: Right Dip Switch Settings  
Switch  
Setting  
all OFF  
Meaning  
BAUD RATE  
set baud rate to 2400  
DATA FORMAT OFF  
odd/even parity  
OFF  
ON  
disable parity checking  
use 2 stop bits  
ON  
use 8 bits per character  
HANDSHAKE  
OFF  
disable XON/XOFF protocol  
10112341 Rev 002  
Page 39 of 39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RM3 SERIES BACKREFLECTION METER  
USER’S MANUAL  
Programming Examples  
RS232 Interface Program Example  
The following program can be used to connect the RM meter with an external device through  
the RS232 serial interface.  
100 CLOSE #1  
' Initialize the RS232C port  
' Go to Backreflection mode  
110 OPEN "COM1:300,N,8,2" AS #1  
200 PRINT #1,"B";  
210 INPUT #1,IN$  
220 IF IN$ = "Y" THEN GOTO 250 ' Check if the command was accepted  
230 PRINT "WILL NOT DISPLAY BR"  
240 STOP  
250 FOR I=1 TO 5  
measurements  
' Read and display five BR  
260 PRINT #1,"G";  
270 INPUT #1,IN$  
280 PRINT IN$  
290 NEXT I  
300 PRINT #1,"A";  
310 INPUT #1,IN$  
' Go to Power mode  
320 IF IN$ = "Y" THEN GOTO 350 ' Check if the command was accepted  
330 PRINT "WILL NOT DISPLAY POWER"  
340 STOP  
350 FOR I=1 TO 5  
measurements  
' Read, display five power  
360 PRINT #1,"G";  
370 INPUT #1,IN$  
380 PRINT IN$  
390 NEXT I  
400 CLOSE #1  
410 END  
10112341 Rev 002  
Page 40 of 39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RM3 SERIES BACKREFLECTION METER  
USER’S MANUAL  
Customized Features and Test Data  
10112341 Rev 002  
Page 41 of 39  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Insignia Toaster NS TWSS2 User Manual
Intel Computer Hardware U3 1L User Manual
JVC Camcorder Accessories LYT2232 002A User Manual
JVC MP3 Player RC EX30 User Manual
Kalorik Refrigerator USK WCL 32963 User Manual
Karcher Pressure Washer G3200XH User Manual
Kenmore Oven 318201516 User Manual
Keys Fitness Home Gym ST VDB User Manual
KitchenAid Double Oven Double Oven User Manual
Knoll Stereo Amplifier MA215E User Manual