User and Service Guide
Publication number 01141-97002
July 2004
For Safety and Regulatory information, see the pages at the end
of this book.
©
Copyright Agilent Technologies 2000-2004
All Rights Reserved
Agilent Technologies 1141A
Differential Probe and 1142A
Probe Control and Power
Module
Download from Www.Somanuals.com. All Manuals Search And Download.
Contents
Accessories Supplied
6
Recommended Test Equipment
9
Calibration Test Record 39
Probe Adjustment 40
Attenuator Adapter Adjustment 49
3 Service
Introduction 54
Performance Specifications and Characteristics 55
General Characteristics 58
3
Download from Www.Somanuals.com. All Manuals Search And Download.
Contents
Differential Probe 60
Probe Control and Power Module Troubleshooting 64
Replaceable Parts 71
Parts List 71
Ordering Information 71
Direct Mail Order System 71
Manufacturers’ Codes 72
Exploded View 73
4
Download from Www.Somanuals.com. All Manuals Search And Download.
Operating the Probe
Accessories Supplied
Introduction
This chapter shows you how to connect and operate the 1141A
Differential Probe and 1142A Probe Control and Power Module as a
differential probe system.
Accessories Supplied
The following items are supplied as part of the 1141A/1142A probe
system. Item numbers refer to the numbers in Figure 1-1 on page -7
and Figure 1-2 on page -8. Those without item numbers are supplied
but not shown in figures. See the Replaceable Parts List for parts not
listed below.
Item
1
Description
Qty
1
Part Number
Differential Probe
2
10x Attenuator Adapter
100x Attenuator
1
5063-2144
5063-2145
5063-2146
5959-9334
1400-1422
5061-6162
01141-68702
01141-66504
8710-1961
1251-5943
1142A
3
1
4
ac Coupling Adapter
Two-inch Extension Leads (package 5)
Mini Grabbers
1
5
1
6
2
7
Five-inch Ground Lead
Shielded Signal Lead
Test Board
1
8
1
9
1
10
11
Flat-blade Alignment Tool
Circuit Connection Posts (strip of 20)
Probe Control and Power Module
Power Cord
1
1
1
1
see parts list
Carrying case
User and Service Manual
1
6
Download from Www.Somanuals.com. All Manuals Search And Download.
Operating the Probe
Accessories Available
Figure 1-2
1141A Miscellaneous Accessories
Figure 1-3
1142A Probe Control and Power Module
Accessories Available
The following accessories can be ordered.
•
•
5959-9335 Long Extension Lead (5.5 inch/14 cm), package of 5
5090-4833 Mini grabber for SMT, package of 20
8
Download from Www.Somanuals.com. All Manuals Search And Download.
Operating the Probe
To inspect the probe
Inspect the shipping container for damage. If the shipping container or cushioning
material is damaged, it should be kept until the contents of the shipment have been
checked for completeness and the instrument had been checked mechanically and
electrically. Accessories supplied with the instrument are listed in see “Accessories
Supplied” on page 6 of this manual.
If the contents are incomplete, if there is mechanical damage or defect, or if the
instrument does not pass calibration tests, notify the nearest Agilent Technologies office.
If the shipping container is damaged, or the cushioning materials show sign of stress,
notify the carrier as well as the nearest Agilent Technologies office. Keep the shipping
materials for the carrier’s inspection. The office will arrange for repair or replacement
at Agilent Technologies’ option without waiting for a claim settlement.
Using the probe with other instruments
The 1141A/1142A probe system can be used with other instruments as well as
oscilloscopes. You can use it with a spectrum analyzer or frequency counter, or any
instrument with an input that can be terminated with 50 Ω.
If you are going to use the probe system with an instrument other than an oscilloscope,
you may need to set up the probe with an oscilloscope first. This will allow you to select
coupling and reject modes, and set offset, so the output of the probe is compatible with
signal requirements of the other instrument.
Recommended Test Equipment
The following table is a list of the test equipment required to test calibration, make
adjustments, and troubleshoot this instrument. The table indicates the critical
specificationsofthetest equipmentandfor whichproceduretheequipmentisnecessary.
Equipment other than the recommended model may be used if it satisfies the critical
specifications listed in the table.
Recommended Test Equipment
Recommended
Agilent
Equipment Required Critical Specifications
Model/Part
Use
P
Signal Generator
Function Generator
Oscilloscope
1 to 200 MHz, 400 m V
rms
8648A
2.5 to 500 kHz, 600 m V to 16 V
33120A
A
p-p
p-p
>400 MHz bandwidth, 1 mV/div sensitivity 54830B
100 mV to 7 Vdc E3632A
P,A
P
dc source
9
Download from Www.Somanuals.com. All Manuals Search And Download.
Operating the Probe
Recommended Test Equipment
Recommended
Agilent
Equipment Required Critical Specifications
Model/Part
Use
A
Test Board
Load
No substitute (supplied accessory)
01141-66504
BNC Feedthrough, 50 Ω
Pasternack
P
Enterprises PE6008-
50 or Huber+Suhner
22543742
Cables (2)
Cable
BNC, 50 Ω 36-inch
10503A
P,A
A
BNC, 50 Ω 9-inch
10502A
Cable
Type N (m) 24-inch
11500B
P
Adapter
Type N (m) to BNC (f)
Type N (f) to BNC (m)
BNC (f) to dual banana (m)
Small flat blade (supplied accessory)
1250-0780
1250-0077
1251-2277
A
Adapter
P
Adapters (2)
Alignment tool
P
8710-1961/
A
Sprague-Goodman
part number GTT-5G
P=Calibration Tests, A=Adjustments
Setting up the probe
The following paragraphs cover system preparation and initial adjustments.
Power Requirements
The 1141A/1142A probe system (specifically the 1142A) requires a power source of
either 90 to 132/198 to 264 Vac, 47 to 440 Hz, 25 VA maximum.
C A U T I O N
Before connecting power to this instrument, be sure the line voltage switch on the rear
panel of the instrument is set properly.
Line Voltage Selection
Before applying power, verify the setting of the LINE SELECT switch on the rear panel
of the 1142A. The slide switch can be set to either 115 or 230 V.
10
Download from Www.Somanuals.com. All Manuals Search And Download.
Operating the Probe
Recommended Test Equipment
W A R N I N G
Before connecting this instrument, the protective earth terminal of the instrument must
be connected to the protective conductor of the (Mains) power cord. The Mains plug
must be inserted in a socket outlet provided with a protective each contact. The
protective action must be negated by the use on an extension cord (power cable)
without a protective conductor (grounding). Grounding one conductor of a two-
conductor outlet does not provide an instrument ground.
This instrument is provided with a three-wire power cable. When connected to an
appropriate ac power outlet, this cable grounds the instrument cabinet. The type of
power cable plug shipped with the instrument depends on the country of destination.
The 1142A Power Control and Power Module does not have a power switch. A power
switch is not required because of the low mains power requirement.
Figure 1-4
1142A Rear Panel
Procedure
1 Use the power cord to connect the 1142A to the ac mains.
2 Connect the 1141A probe cable power connector to the PROBE connector on
the rear panel of the 1142A power module.
3 Connect the output of the probe to the input of the oscilloscope.
4 Set the input impedance of the oscilloscope to 50 Ω.
If the oscilloscope does not have a selectable 50 Ω input impedance, connect a 50 Ω BNC
feedthrough termination between the probe output and the input of the oscilloscope.
5 If making an initial equipment setup, continue with the initial adjustment in
the following section.
Initial Adjustment
For a given combination of 1141A Differential Probe and 1142A Probe Control and Power
Module, you may want to adjust the Offset Null and DC Reject Gain. Typically, you need
to make these adjustments only once, before the probe is first used. You can make them
any time to optimize the system. These adjustments do not affect the specifications of
the probe system.
11
Download from Www.Somanuals.com. All Manuals Search And Download.
Operating the Probe
Recommended Test Equipment
•
•
Offset null zeroes the dc level at the output of the probe. The range of adjustment is
about 4 mV.
DC Reject Gain adjusts the gain of the dc reject circuit to accurately null the dc
component of an input signal. The range of adjustment is about 0.5%.
Figure 1-5
1142A Front Panel
Equipment Needed
The following equipment is necessary for initial adjustment.
• 5 V power supply
• DVM that can measure 25 µV
• 50 Ω BNC feedthrough terminator
• 01141-66504 test board
Equipment Setup
Use the following procedure to setup the differential probe system for initial adjustment.
C A U T I O N
Do not exceed 7 V when using the test board for this procedure. If the voltage is too
high, it will cause excessive power dissipation in the 50 Ω termination on the test board.
1 Use the probe setup procedure to set up the probe system.
2 Connect a 50 Ω BNC feedthrough terminator to the output of the probe.
3 Disconnect all accessories from the input of the probe.
4 Connect the DVM to measure the dc output of the probe at the 50 Ω load.
12
Download from Www.Somanuals.com. All Manuals Search And Download.
Operating the Probe
Recommended Test Equipment
5 Set up the 1142A:
a
Set the Local/Remote push button to Local.
b
Under DC Couple, press Zero offset.
6 Set the power supply output to 5 V.
7 Arrange a connection between the power supply and the test board. The
negative terminal of the supply should connect to the shield of the test board
BNC.
If your power supply has standard binding posts, you can connect a banana-to-BNC
adapter to the supply and connect a BNC cable between the supply and the test board.
Adjustment
Warm up the 1141A for 30 minutes before making adjustments.
1 With the 1141A probe inputs unconnected, adjust Offset Null on the 1142A for
a minimum reading on the DVM.
The voltage swing of the adjustment is approximately 4 mV.
2 On the 1142A, under DC Reject, press 5.0 Hz.
3 Read and record the reading on the DVM, _________ mV.
4 Connect the probe to the test board in the position shown below
(signal to + input).
Figure 1-6
Signal to + input
5 AftertheDVMreadingstabilizes, adjustDCRejectGaintothereadingrecorded
in step 3.
With a 5 V supply, the voltage swing is approximately 12.5 mV. With a lower supply, the
voltage swings proportionally less.
Using the Accessories
The 1141A Differential Probe and accessories are designed to provide a variety of ways
to connect to circuitry and make measurements. In the descriptions, any method used
to connect to the probe signal inputs also applied to the adapters. The figure below
shows, in a general way, the use of accessories.
13
Download from Www.Somanuals.com. All Manuals Search And Download.
Operating the Probe
Recommended Test Equipment
Figure 1-7
Basic Accessory Connections
Probe Tips
Probe tips fit into the receptacles in the probe and are held in place with probe tip caps.
If necessary, you can solder the probe tips into a circuit or wires can be soldered to the
tips. If you solder to the probe tips, be careful not to melt the plastic probe tip caps.
N o t e
Because of the close tolerances between the probe tip caps and probe tips, it will be
difficult to separate the probe tips and caps once the probe tips have been soldered.
Ground Leads
The circular end of the ground lead fits over the screw on the top side of the probe.
Extension Leads
The extension leads provide a flexible connection between circuitry and the probe.
•
To provide a male connection to other circuitry, connect the extension lead over the
probe tips.
•
Toprovideafemaleconnection, removetheprobetipcapsandprobetipsandconnect
the extension leads to the probe.
14
Download from Www.Somanuals.com. All Manuals Search And Download.
Operating the Probe
Recommended Test Equipment
N o t e
Use extension leads and similar connection accessories carefully. Extension leads
compromise the high-frequency specifications of the probe. CMRR is particularly
sensitive to unbalanced input parameters.
To prevent pickup of stray fields when you use extension lead, either the ones supplied
with the 1141A or others, dress them carefully as follows:
•
•
Connect the leads at right angles to the circuitry under test.
Keep the leads as parallel as possible before they connect to the probe.
Mini-Grabbers
Mini-grabbers can be attached to the probe or adapter through the extension leads.
1 Remove the probe tip caps and tips.
2 Attach the extension leads to the probe or adapter.
3 Attach the mini-grabbers to the extension leads.
Circuit Connector Posts
These 0.025-inch square posts can be used to connect either directly to the probe or to
the extension leads.
Solder the posts directly into your circuitry or use them to make extension leads that
plug into the inputs of the probe or adapters.
Shielded Signal Leads
The shielded signal leads allow connection to points in a circuit that are up to 10 inches
apart. The leads are shielded to within approximately 1/2 inch of the end of the lead so
they minimize pick-up due to stray fields from adjacent circuitry.
1 Connect the end with the ground connector to the probe pins and ground of
the differential probe or adapter.
2 Connect the free ends of the leads to 0.025-inch square or 0.030-inch round pins
in your circuitry or to the mini-grabbers.
N o t e
Each lead has an input capacitance of approximately 15 pF. This capacitance may limit
the bandwidth of your measurement (depending on the impedance of the circuit). Also,
CMRR may be affected because of slight differences between the input capacitance of
the two leads. CMRR is also affected by differences in impedance between the two
measurements points.
15
Download from Www.Somanuals.com. All Manuals Search And Download.
Operating the Probe
Recommended Test Equipment
Adapters
There are three adapters for use with the differential probe. Two adapters are
attenuators, a 10x and a100x. The other is an ac adapter for blocking dc from the probe
input.
The adapters are installed on the probe after the probe tip caps and probe tips have been
removed. The adapter fastens to the probe using a thumb wheel located on the underside
of the adapter. the figure below shows a good way to hold the probe while attaching the
adapter.
1 Remove the probe tip caps and probe tips, from the probe.
2 Fit the adapter over the end of the probe and rotate the thumb wheel with your
finger until the adapter fits snugly.
A snug fit is important because the ground is maintained through the thumb wheel screw.
A loosely attached adapter compromises the mechanical and electrical integrity of the
combination.
Figure 1-8
Attaching the Adapters
16
Download from Www.Somanuals.com. All Manuals Search And Download.
Operating the Probe
Recommended Test Equipment
Adapter Combinations
The figure below shows the allowed adapter and probe connections. There are two
specific combinations that should not be used.
•
•
Do not attach the ac adapter between an attenuator adapter and the probe.
An attenuator adapter must be terminated by the input resistance of the probe. The
ac adapter isolates the probe input resistance.
Do not cascade two attenuator adapters.
The attenuator adapters are designed to be terminated by the 1 MΩ resistance of the
probe. The input resistance of the attenuator adapter is 9 MΩ for the 10x adapter and
10 MΩ for the 100x adapter.
Figure 1-9
Allowed Adapter Connections
17
Download from Www.Somanuals.com. All Manuals Search And Download.
Operating the Probe
Recommended Test Equipment
Connector Compatibility
The following are general connector characteristics for the probe, adapters, and
accessories.
•
The female connectors on the probe, adapters, and other accessories are designed to
mate with 0.030-inch round or 0.0250-inch square pins.
•
•
The probe, adapter, and extension lead pins are 0.030-inch round.
The strip of circuit connection posts provided as an accessory has 0.025-inch square
pins.
•
•
The mini-grabber has a 0.25-inch square pin.
The ground connection at the end of the probe and adapters (where the adapters
fasten) accepts an M3 metric screw.
Test Board
The primary use of the test board is to apply test and calibration signals to the input of
the probe or adapters. Specific use of the test board is covered wherever it applies.
Grounding
Grounding is very important when probing circuitry. Improper grounding can increase
the common mode signal level. This reduces the effectiveness of the differential probe.
The mechanical connections at the input of the probe are ground for probe signals. The
screw where the ground lead attaches (see figure 1-7) fastens to this ground. Also, the
attenuator and ac adapter fasten to this ground through the screw connection and the
ground is carried through each adapter to its front.
Figure 1-10
Probe Grounds
Coupling Functions
There are three methods for blocking or compensating for the dc component of a signal.
Each has specific advantages.
18
Download from Www.Somanuals.com. All Manuals Search And Download.
Operating the Probe
Recommended Test Equipment
Probe System Coupling Functions
dc offset
dc Reject
ac Coupling
Adapter
dc Blocked
20 V
200 V
500 V
20 V
200 V
500 V
20 V
500 V
500 V
Probe alone
Probe with 10x
Probe with 100x
Adjust offset to put
signal on screen
Select DC Reject
low-frequency
corner
Attach ac Coupling
adapter to
differential probe
Set-up needed
1
1
Remote Control?
Yes
Yes
No
CMRR
No
No
No
Yes
degradation?
Low-frequency
degradation?
Yes
Yes
1
Isolated external dc reference and control signals are needed
dc Reject
dc Reject is the best method of eliminating the dc component of a signal when dc is not
a factor in the measurement.
The key characteristics are:
•
•
•
•
The low-frequency component (from dc to the selected corner frequency) is
automatically nulled by the dc reject circuitry.
Probe CMRR specifications are not compromised as happens when the ac coupling
adapter is uses.
There is a selectable low frequency corner with -3dB points at 0.05 Hz, 0.5 Hz,
or 5.0 Hz.
Thevoltagerejectrangeis 20Vwiththeprobealone, 200Vwiththe10xattenuator,
and 500 V with the 100x attenuator.
19
Download from Www.Somanuals.com. All Manuals Search And Download.
Operating the Probe
Recommended Test Equipment
To use dc reject:
1 Remove the ac adapter if it is installed.
2 On the front panel of the 1142A, press Local.
3 Under DC Reject on the front panel, press 5.0 Hz or 0.5 Hz individually, or
5.0 Hz and 0.5 Hz simultaneously to get 0.05 Hz.
Within the frequency and voltage characteristics noted elsewhere in this manual, low
frequencies are nulled from the input signal.
Offset
Offset is the best method to use when the low-frequency corners associated with dc
reject and the ac adapter interfere with the measurement.
The key characteristics are:
•
•
•
The user manually null the dc component with the offset adjustment.
Offset is dc coupled so there is no low frequency roll-off.
Probe CMRR specifications are not compromised as happens when the ac coupling
adapter is used.
•
Thevoltageoffsetrange is 20V with theprobealone, 200V with the10xattenuator,
and 500 V with the 100x attenuator. (With the 100x attenuator, the offset range is
restricted by the maximum input voltage rating rather than the operating range of
the offset).
To use offset:
1 Remove the ac adapter if it is installed.
2 On the front panel of the 1142A, press Local and Variable offset.
3 Adjust the Coarse and Fine Variable Offset until the signal is displayed on the
screen of the oscilloscope.
ac Adapter
The ac adapter must be used when the dc component of the signal exceeds the operating
range of the dc reject or offset methods. The ac adapter block the dc and low frequency
component of the input by forming a high pass filter with the input impedance of the
probe or adapter.
The key characteristics are:
•
The ac adapter safely blocks 200 Vdc when attached directly to the probe or
500 Vdc when attached to a 10x or 100x adapter.
•
The probe and adapters have different input impedances, so they have different low
frequency corners with the ac adapter. When the ac adapter is directly on the probe
the -3dB corner is 15 Hz. When the ac adapter is on an attenuator the corner is 1.5 Hz.
•
The low-frequency CMRR when using the ac adapter is not as good as when using the
probe alone or the probe with a 10x or 100x adapter.
C A U T I O N
If you measure a node having a high dc potential, the blocking capacitors in the ac
adapter will charge to that potential. After making such measurements, discharge the
capacitors by grounding both inputs of the ac adapter. This will prevent damage by a
high voltage discharge into sensitive circuitry when the next measurement is made.
20
Download from Www.Somanuals.com. All Manuals Search And Download.
Operating the Probe
Recommended Test Equipment
To use ac coupling:
1 Attach the ac coupling adapter to the input of the probe or the input of the
attenuator adapter.
2 On the 1142A, press Local and Zero offset.
Remote operation
For automatic test applications, the coupling and offset functions provided by the 1142A
Probe Control and Power Module can be remotely controlled through a connector on the
rear panel of the module. The connection is through a standard 9-pin female
D-subminiature connector. This style is the same as that used on some personal
computer monitor cables, which provides an economical way to connect the 1142A to
the controller interface on an automatic test system.
The following table gives the connections.
Remote Input Connections
Pin
1
Function
Pin
6
Function
Connector
Function Select 1 (A1R)
Digital common
N.C.
Function Select 0 (A0R)
2
7
N.C.
3
8
N.C.
4
External offset common
Shield
9
External offset
5
N O T E
To minimize dc offset errors and potential noise coupling, electrically isolate all
connections between the Remote Input connector and the controlling system.
21
Download from Www.Somanuals.com. All Manuals Search And Download.
Operating the Probe
Recommended Test Equipment
Function Select
The easiest way to control the function select lines is contact closures between the lines
and Digital common (pin 2) of the remote input connector. (TTL compatible control
signals can be used; but to avoid problems with ground loops, they must be electrically
isolated.) The following truth table shows the functions provided by the function select
lines. For the Remote Inputs, “0” represents a closure and “1” represents an open circuit.
Remote Functions Select Truth Table
Remote Input
Function
A1R
(Pin 1)
A0R
(Pin 6)
0
0
1
1
0
1
0
1
0.05 Hz Reject
0.5 Hz Reject
5 Hz Reject
DC Couple
Variable Offset
The remote variable offset can be used when the dc couple function is remotely selected.
The offset voltage must be referenced to the External offset common (pin 4) of the
remote input connector. It must be electrically isolated from the controlling system. The
following table shows the offset range and remote offset requirements for probe and
adapter combinations.
Remote Offset Input Requirements
Offset Range
Remote
Requirements
Probe alone
20 V
200 V
500 V
10 V
10 V
2.5 V
Probe with 10x adapter
Probe with 100x adapter
Cleaning Requirements
If the instrument requires cleaning: (1) Remove power from the instrument. (2) Clean
the external surfaces of the instrument with a soft cloth dampened with a mixture of
mild detergent and water. (3) Make sure that the instrument is completely dry before
reconnecting it to a power source.
22
Download from Www.Somanuals.com. All Manuals Search And Download.
Operating the Probe
Recommended Test Equipment
Differential Amplifiers and CMRR
The 1141A Differential Probe is a high-impedance differential amplifier. A characteristic
of differential amplifiers is the ability to reject signals that are common to the two inputs.
The common mode rejection ratio (CMRR) is the measurement of this ability. It is
expressed as the ratio between the amplitudes of the common mode and differential
signals which product equal outputs. For example, if a common mode signal of 1 V and
differential signal of 1 mV both produce outputs of 1 mV, the CMRR is 1000:1.
The ability to reject common mode signals is dependent on the balance designed into
the differential amplifier. At higher frequencies it becomes harder to balance circuit
parasitics and parameters of devices so CMRR degrades as frequency increases. Also,
stray coupling increases with frequency and coupling may vary between the two
differential paths.
The CMRR of the 1141A Differential Probe is specified at the input of the probe and
cannot be affected expect by adjustments in the probe. However, the way the probe is
connected into the circuitry being tested can have a big influence in the overall result of
the measurement, especially at high frequencies.
The following things can affect the effective CMRR of a test setup:
•
The connection to the circuit under test.
The method used to connect the probe is important because it involves the symmetry
of the differential input circuitry. For example, using different lengths of wire to
connect the circuit to the two probe inputs unbalances the inductance and
capacitance at the inputs. The effective CMRR will be reduced, especially at high
frequencies. Additionally, coupling from adjacent circuitry will be less balanced.
•
The impedance of the source.
This is another instance where the symmetry of the differential circuit is important.
The impedance of the source forms a network with the input impedance of the source
forms a network with the input impedance of the connections and the probe. This
network determines the frequency response for the measurement. If each side of the
differential source has a different impedance, the frequency response of each side
will be different and the unbalance is reflected in a reduced CMRR. Of course, lower
source impedances have less effect on the frequency response of the measurement.
•
•
The ground connection.
A poorly located ground connection allows ground loops to add to the common mode
signal.
Frequency.
Frequency is the most important factor in CMRR only because all of the factors
mentioned above are frequency dependent. The unbalances of capacitance and
inductance are more important as frequency increases. Therefore, good high-
frequency practice is important when using a high impedance differential probe.
On the other hand, if the differential probe is ac coupled to the circuit under test (the
ac adapter is being used) the CMRR will be degraded below a certain frequency; the
lower the frequency the worse the CMRR. This is because unbalance in the series
capacitances of the ac coupler becomes more significant the lower the frequency.
23
Download from Www.Somanuals.com. All Manuals Search And Download.
Operating the Probe
Recommended Test Equipment
24
Download from Www.Somanuals.com. All Manuals Search And Download.
Calibration Tests and Adjustment
Equipment Required
Introduction
This chapter is divided into two sections. The first section gives
calibration tests and the second adjustment procedures for the 1141A
Equipment Required
A complete list of equipment required for the calibration tests and adjustments is listed
in “Recommended Test Equipment” on page 9. Equipment required for individual
proceduresislistedatthe procedure. Anyequipmentsatisfying thecriticalspecifications
listed may be substituted for the recommended model.
The Test Board
The test board is a supplied accessory for use during calibration tests and adjustments
to connect signals to the differential probe (with or without adapters). A BNC connector
connects the test board to a cable from the signal generator. The board includes a 50 Ω
termination (two 100 Ω resistors).
C A U T I O N
The power rating of the 50 Ω termination is 1.0 W. Keep the signal input below 7 Vdc
or rms to avoid degrading the termination.
Once the probe tip caps and probe tips have been removed, the probe can be connected
to the test board in one of three ways, as shown in the figure below. At each position of
the probe, probe inputs are connected to a different combination of signal and ground.
A separate terminal on the test board connects the ground of the probe to the signal
ground.
Figure 2-1
Test Board Showing Probe Positions
26
Download from Www.Somanuals.com. All Manuals Search And Download.
Calibration Tests and Adjustment
The Test Board
Calibration Tests
These procedures test the probe’s electrical performance using applicable specifications
given in “Performance Specifications and Characteristics” on page 55 as performance
standards. Specifications applicable to individual tests are noted at the test for reference.
Testing Interval
The calibration testing procedures may be performed for incoming inspection of the
instrument and should be performed periodically thereafter to ensure and maintain peak
performance. The recommended test interval is yearly or every 2,000 hours of operation.
Amount of use, environmental conditions, and the user’s experience concerning need
for testing will contribute to verification requirements.
Calibration Test Record
The results of the calibration tests may be tabulated in the Calibration Test Record
provided at the end of the calibration tests. The Calibration Test Record listed the
calibration tests and provides an area to mark test results. The results recorded in the
Calibration Test Record during initial inspection may be used for later comparisons of
the tests during periodic maintenance, troubleshooting, and after repairs or adjustments.
Calibration Test Procedures
Procedures may be done individually or in any order.
N O T E
Allow the instrument to warm up for at least 30 minutes prior to beginning calibration
tests.
27
Download from Www.Somanuals.com. All Manuals Search And Download.
Calibration Tests and Adjustment
dc Gain Accuracy
dc Gain Accuracy
This test checks the dc gain accuracy of the differential probe and the dc accuracy of
the differential probe with attenuator and adapters.
Specification: Probe alone, 2%; with attenuator adapter, 4%
Equipment Required
Recommended
Equipment
Required
Agilent
Critical Specifications
Model/Part
ac/dc Calibrator
or
dc Power Supply
100 mV to 7 V
E3632A
100 mV to 7 V
DVM
Load
0.5% accuracy, 10 µV resolution 3458A or
E34401A
BNC Feedthrough, 50 Ω
Pasternack
Enterprises
PE6008-50 or
Huber+Suhner
22543742
Cables (2)
BNC 50 Ω
10503A
Adapters (2)
Test Board
BNC (f) to dual banana (m)
No substitute
1251-2277
01141-66514
Probe Gain Test Procedure
1 Connect the probe and test equipment as shown in figure 2-2.
28
Download from Www.Somanuals.com. All Manuals Search And Download.
Calibration Tests and Adjustment
dc Gain Accuracy
Figure 2-2
2 Set up the 1142A probe control and power module as follows:
b
Under DC Couple, press the Zero offset button.
3 With the 1141A Probe Amp disconnected from the test PCA, adjust the Offset
Null control on the 1142A until the DVM reads 0Vdc.
measurements in this test.
Figure 2-4 on page -31.
5 Adjust the DC Source to output 100 mV (nom.)
6 Record the V measurement from the top DVM in figure 2-2.
in1
7 Record the V
measurement from the bottom DVM in figure 2-2.
out1
8 Connect the Probe Amp to measure a negative voltage as shown in figure 2-3.
9 Record the V measurement from the bottom DVM in figure 2-2
out2
29
Download from Www.Somanuals.com. All Manuals Search And Download.
Calibration Tests and Adjustment
dc Gain Accuracy
Figure 2-3
Signal to - input
10 Calculate probe gain as
(Vout – Vout
)
∆Vout
-------------
∆Vin
1
2
= ---------------------------------
2 × Vin
1
Record the result of this calculation in the “Calibration Test Record” on page 39.
To pass this test, the probe gain = 0.98 to 1.02
N O T E
N O T E
Failure of the gain accuracy test can be caused by mis-adjustment of the probe. Perform
the Probe Adjustment procedure in the Adjustments section later in this chapter and
retest.
10x Attenuator Accuracy Test
If the gain test for the probe fails, the 10x Attenuator Accuracy Test will fail or the
results will be poor. Do not continue until the probe passes the dc gain test.
1 Disconnect the probe from the test board and connect the 10x attenuator
adapter to the probe.
2 Carefully connect the input of the probe/attenuator to the test board in the
position shown in the figure below (signal to + input).
30
Download from Www.Somanuals.com. All Manuals Search And Download.
Calibration Tests and Adjustment
dc Gain Accuracy
Figure 2-4
Signal to + input
3 Set the dc calibrator output 3 V dc.
4 Record the V measurement from the top DVM in figure 2-2.
in1
5 Record the V
measurement from the bottom DVM in figure 2-2
out1
6 Carefully connect the input of the probe/attenuator to the test board in the
position shown in the figure below (signal to - input).
Figure 2-5
Signal to - input
7 Record the V
measurement from the bottom DVM in figure 2-2.
out2
8 Calculate the 10x attenuator gain as
(Vout – Vout
)
∆Vout
-------------
∆Vin
1
2
= ---------------------------------
2 × Vin
1
Record the result of this calculation in the “Calibration Test Record” on page 39.
N O T E
Failure of the accuracy test for the 10x attenuator can be caused by mis-adjustment of
the low-frequency CMRR (LF CMRR) adjustment. Perform the Attenuator Adapter
Adjustment procedure in the Adjustments section later in this chapter then retest the
attenuator adapter. If if continues to fail, repair is necessary.
31
Download from Www.Somanuals.com. All Manuals Search And Download.
Calibration Tests and Adjustment
dc Gain Accuracy
100x Attenuator Accuracy Test
N O T E
If the gain test for the probe fails, it will be reflected in the test for the 100x attenuator
adapter. Do not continue until the probe passes the gain test.
1 Disconnect the probe/attenuator from the test board. Remove the 10x
attenuator adapter from the probe and connect the 100x attenuator adapter.
2 Carefully connect the input of the probe/attenuator to the test board in the
position shown in the figure below (signal to + input).
Figure 2-6
Signal to + input
C A U T I O N
Avoidexcessivepowerdissipationintheterminationonthetestboard. Keepthevoltage
input at or below 7 Vdc.
3 Set the dc source output to 7 Vdc.
4 Record the V measurement from the top DVM in figure 2-2.
in1
5 Record the V
measurement from the bottom DVM in figure 2-2
out1
32
Download from Www.Somanuals.com. All Manuals Search And Download.
Calibration Tests and Adjustment
dc Gain Accuracy
6 Carefully connect the input of the probe/attenuator to the test board in the
position shown in the figure below (signal to - input).
Figure 2-7
Signal to - input
7 Record the V
measurement from the bottom DVM in figure 2-2.
out2
8 Calculate the 100x attenuator gain as
(Vout – Vout
)
∆Vout
-------------
∆Vin
1
2
= ---------------------------------
2 × Vin
1
Record the result of this calculation in the “Calibration Test Record” on page 39
N O T E
Failure of the accuracy test for the 100x attenuator can be caused by mis-adjustment
of the low-frequency CMRR (LF CMRR) adjustment. Perform the Attenuator Adapter
Adjustment procedure in the Adjustments section later in this chapter then retest the
attenuator adapter. If if continues to fail, repair is necessary.
33
Download from Www.Somanuals.com. All Manuals Search And Download.
Calibration Tests and Adjustment
Bandwidth
Bandwidth
This test checks the high-frequency response of the 1141A Differential Probe. The
bandwidth of the oscilloscope is characterized first so it is not a factor in the
measurement.
Specification (-3dB, dc coupled): dc to 200 MHz
Equipment Required
Equipment
Required
Critical Specifications
Recommended
Model/Part
Oscilloscope
Signal Generator
Test Board
Cable
400 MHz bandwidth
200 MHz at ≈ 230 mVrms
No substitute
54830A
8648A
01141-66504
11500B
Type N (m) 24-inch
Type N (f) to BNC (m)
Adapter
1250-0077
Procedure
This test depends on the accuracy of the termination on the test board and the
termination in the oscilloscope. Both should be with 1%.
1 With the N cable and N-to-BNC adapter, connect the signal generator to the
oscilloscope channel 1 input.
2 Set the signal generator for 200 MHz at 0 dBm (about 224 mVrms).
3 Set the 1142A front panel switches to Local and Zero offset.
4 On the oscilloscope press AUTOSCALE, then set the following parameters.
Menu
Selection
Setting
TIMEBASE
CHAN 1
(time/div)
2 ns/div
(sensitivity)
(input R)
100 mV/div
50 Ω DC
ACQUISITION
Sampling Mode
Memory Depth
Sample Rate
Averaging
Real Time
Automatic
Automatic
Enabled
32
# of avg
5 The signal on screen should be about six divisions amplitude.
Measure the peak-to-peak voltage of the channel 1 signal and record the
reading.
V
(1) = _____________ mV
p-p
34
Download from Www.Somanuals.com. All Manuals Search And Download.
Calibration Tests and Adjustment
Bandwidth
6 Reconfigure the equipment.
a
b
c
Disconnect the signal generator cable from the oscilloscope input and connect it to
the test board.
Connect the output of the differential probe to the channel 1 input of the
oscilloscope.
Carefully connect the input of the probe to the test board in the position shown in
the figure below (signal to + input).
Figure 2-8
Signal to + input
7 Record the V reading on the oscilloscope.
p-p
V
(1) = _____________ mV
p-p
8 Divide the reading from step 6 by the reading from step 4.
Answer from step 6
Answer from step 4
----------------------------------------------
= ____________
Record the result in the Calibration Test Record.
The result should be 0.707 or greater, indicating a probe bandwidth of 200 MHz or more.
The bandwidth can be checked at other signal levels. Change the signal generator output
level and oscilloscope V/div range proportionally.
N O T E
Failure of the bandwidth test can be caused by mis-adjustment of the probe. Perform
the Probe Adjustment procedure in the Adjustments section later in this chapter.
35
Download from Www.Somanuals.com. All Manuals Search And Download.
Calibration Tests and Adjustment
CMRR Test
CMRR Test
This test checks the CMRR at 1 MHz and 100 MHz.
Specification 3000:1 at 1 MHz, 10:1 at 100 MHz
Equipment Required
Recommended
Agilent
Model/Part
Equipment
Required
Critical Specifications
Oscilloscope
Signal Generator
Test Board
Cable
400 MHz bandwidth at 1 mV/div 54830B
1-100 MHz at = 400 mVrms
No substitute
8648A
01141-66504
11500B
Type N (m) 24-inch
Type N (f) to BNC (m)
Adapter
1250-0077
Procedure
1 Connect the probe power connector to the PROBE connection on the rear of
the 1142A Probe Control and Power Module.
2 Set the 1142A front panel switches to Local and Zero offset.
3 Connect the probe output to the oscilloscope channel 1 input.
4 Connect the input of the probe to the test board in the position shown in the
figure below (signal to + input).
Figure 2-9
Signal to + input
5 Connect the signal generator to the test board.
6 Set the signal generator for 1 MHz at 385 mV
(1 V ).
p-p
rms
36
Download from Www.Somanuals.com. All Manuals Search And Download.
Calibration Tests and Adjustment
CMRR Test
7 On the oscilloscope, press AUTOSCALE and set the following parameters.
Menu
Selection
Setting
TIMEBASE
CHAN 1
(time/div)
500 ns/div
(sensitivity)
(input R)
200 mV/div
50 Ω DC
ACQUISITION
Sampling Mode
Memory Depth
Sample Rate
Averaging
Real Time
Automatic
Automatic
Enabled
32
# of avg
8 On the oscilloscope, measure the peak-to-peak voltage of the channel 1 signal,
then V P-P, then press 1) and record the reading.
V
(1) = _____________ mV
p-p
9 Connect the input of the probe to the test board in the position shown in the
figure below (signal to both inputs).
Figure 2-10
Signal to both input
10 Set the sensitivity to 1 mV/div.
11 After the measurement settles (averaging is complete), record the V P-P
reading.
V
(2) = _____________ mV
p-p
12 Disconnect the probe amp from the test board and measure V
on
noise pp
channel 1.
13 Calculate the CMRR result as follows
V
pp1
CMRR = ------------------------------------
V
– Vnoisepp
pp2
14 The result in step 12 should be ≥ 3000, representing a CMRR of 3000:1 or more.
Record the CMRR in the Calibration Test Record.
37
Download from Www.Somanuals.com. All Manuals Search And Download.
Calibration Tests and Adjustment
CMRR Test
15 Connect the input of the probe to the test board in the position shown in the
figure below (signal to + input).
Figure 2-11
Signal to + input
16 Set the signal generator for 100 MHz at 0.0 dBm (about 224 mV , 632 mV ).
rms
p-p
17 Set the oscilloscope to channel 1 and change the horizontal scale to 5 ns/div.
18 After the measurement settles (averaging is complete), note the V P-P reading.
V
(1) = _____________ mV
p-p
19 Connect the input of the probe to the test board in the position shown in the
figure below (signal to both inputs).
Figure 2-12
Signal to both input
20 Set the channel 1 sensitivity to 10 mV/div.
21 After the measurement settles (averaging is complete), not the V P-P reading.
V
(2) = _____________ mV
p-p
22 Disconnect the probe amp from the test board and measure V
on
noise pp
channel 1.
23 Calculate the CMRR result as follows
V
pp1
CMRR = ------------------------------------
V
– Vnoisepp
pp2
24 The result in step 21 should be ≥ 10, representing a CMRR of 10:1 or more.
Record the CMRR in the Calibration Test Record.
38
Download from Www.Somanuals.com. All Manuals Search And Download.
Calibration Tests and Adjustment
Calibration Test Record
Calibration Test Record
1141A/1142A Differential Probe
Tested by_________________________
Work Order No.____________________
Date____________________
Serial No. ______________________________
Recommended Test Interval - 1 Year/2000 hours
Recommended next testing_________________
Temperature_____________
Test
Limits
Results
dc Gain Accuracy
Probe
Only
+0.98 mV to +1.02 mV
_____________
10x
+0.096 mV to +0.104 mV
+0.0096 mV to 0.0104 mV
>0.707 at 200 MHz
_____________
_____________
_____________
_____________
100x
Bandwidth
CMRR
1 MHz
≥3000:1
≥10:1
100 MHz
39
Download from Www.Somanuals.com. All Manuals Search And Download.
Calibration Tests and Adjustment
Probe Adjustment
Adjustments
This section provides adjustment procedures for the 1141A Differential Probe and
attenuator adapters. There are no service adjustments for the 1142A Probe Control
Module.
Adjustment Interval
None of the adjustment procedures that follow should be considered for a routine
maintenance plan. The differential probe and attenuator adapters should be adjusted
under conditions specified at the beginning of the respective procedures.
N O T E
Warm up the instrument for 30 minutes before starting adjustment procedures.
Probe Adjustment
This procedure adjust the high-frequency and low-frequency paths on the 1141A
Differential Probe.
N O T E
Do not perform this procedure as a part of routine maintenance. Perform the procedure
only if the probe does not meet specifications or has been repaired.
C A U T I O N
You are going to remove the covers of the probe, so the assembly inside will be exposed
while under power. The PC assembly will be electrically and mechanically vulnerable.
Do these adjustment procedures in an ESD-safe area.
Avoid inadvertent contact between the powered assembly and nearby tools and
equipment.
Avoid mechanical damage by carefully handling the exposed assembly and cables.
40
Download from Www.Somanuals.com. All Manuals Search And Download.
Calibration Tests and Adjustment
Probe Adjustment
The following equipment is required for this procedure. Procedures are based on the
model or part number recommended.
Equipment Required
Equipment Required
Critical Specifications
Recommended
Agilent
Model/Part
Function Generator
Oscilloscope
Signal Generator
Test Board
2.5 kHz, 1 V
33120A
p-p
300 MHz at 1 mV/div
200 MHz at ≈ 300 mVrms
No substitute
54830B
8648A
01141-66504
10503A
Cables (2)
BNC, 50 Ω
Adapter
Type N (m) to BNC (f)
1250-0780
Probe Preparation
The probe cover must be removed before adjustment. Drift due to temperature
differences with and without covers is negligible.
1 Remove the probe tip caps and probe tips.
2 Loosen the probe clamp ring at the cable end of the probe (1/4 turn counter-
clockwise) and slide it down the cable.
3 Remove the bottom cover.
a
At the cable end of the probe, separate the covers about centimeter (1/2 inch).
b
Slide the bottom cover toward the cable end of the probe until the locator pins at
the probe input clear the holes. Then, remove the cover.
N O T E
Note the position of the ground block at the input end of the probe. The ground block
is held, through the PC assembly, by the grounding screw on the top of the probe. The
ground block must be reinstalled on the PC assembly after the top cover is removed.
Handle the PC assembly by the edges of the PC board.
4 Remove the ground connection screw on the top of the probe.
The ground block will become free.
At the cable end, the PC board fits over pins inside the top cover.
5 Lift the board off of the pins in the cover and slide it in the direction of the
cable until the input connectors clear the front of the probe.
41
Download from Www.Somanuals.com. All Manuals Search And Download.
Calibration Tests and Adjustment
Adjustment Procedure
6 As shown in the figure below, use the grounding screw to reinstall the ground
block on the PC assembly.
Figure 2-13
Attaching Ground Block to Probe PC Assembly
The ground block provides a mechanical and electrical connection when the probe PC
assembly is connected to the test board.
7 Connect the probe power connector to the PROBE connection on the rear of
the 1142A Probe Control and Power Module.
8 Connect the mains power to the 1142A.
9 Set the 1142A front panel switches to Local and Zero offset.
Adjustment Procedure
Unless specified elsewhere, the procedures must be followed in the order given.
The only adjustment which may be done separately is HF COMP, the high-frequency
compensation.
HF Gain and HF CMRR
This adjustment sequence adjusts the HF Gain for unity gain at 500 kHz and the HF
CMRR for minimum with a 500 kHz common mode signal.
1 Set up the function generator.
•
•
•
Sine wave
500 kHz
600 mV
p-p
2 Use BNC cables to connect the function generator to the oscilloscope.
•
Generator OUTPUT to oscilloscope channel 1 input
•
Generator TRIG OUTPUT to oscilloscope EXT TRIG
42
Download from Www.Somanuals.com. All Manuals Search And Download.
Calibration Tests and Adjustment
Adjustment Procedure
3 On the oscilloscope, then press CLEAR DISPLAY. Press AUTOSCALE, then set
up the following parameters.
Menu
Selection
Setting
TIMEBASE
CHAN 1
(time/div)
500 ns/div
(sensitivity)
(input R)
100 mV/div
50 Ω DC
TRIG
(mode)
source
level
trg’d
EXT
1.00000 V
ACQUISITION
Sampling Mode
Memory Depth
Sample Rate
Averaging
Real Time
Automatic
Automatic
Enabled
32
# of avg
4 On the oscilloscope, measure the peak-to-peak voltage of the channel 1 signal
and record the reading.
V
(1) = _____________ mV
p-p
5 Disconnect the BNC cable from the channel 1 input and connect it to the BNC
connector on the test board.
6 Connect the output of the probe to the channel 1 input.
7 Carefully connect the input of the probe to the test board in the position shown
in the figure below (signal to + input).
Figure 2-14
Signal to + input
43
Download from Www.Somanuals.com. All Manuals Search And Download.
Calibration Tests and Adjustment
Adjustment Procedure
8 Center adjustment R11, HF CMRR (see following figure).
Figure 2-15
R11, HF CMRR Adjustment
9 Adjust R9, HF GAIN so the V (1) measurement is the same as in step 4, within
p-p
0.5%.
Make the adjustment slowly so the oscilloscope display has time to react to signal
averaging. Press CLEAR DISPLAY occasionally to restart averaging, which gives a
quicker indication of changes.
10 Carefully connect the probe to the test board in the positionshown in the figure
below (signal to both inputs).
Figure 2-16
Signal to both inputs
11 Set the function generator output to 1.0 V
.
p-p
12 On the oscilloscope, set the channel 1 sensitivity to 1.00 mV/div.
13 Adjust R11 for minimum signal amplitude as shown in V reading for
p-p
channel 1. Adjust R11 slowly and use CLEAR DISPLAY frequently to restart
averaging.
Low Frequency Response and CMRR
This adjustment sequence continues from the HF Gain and HF CMRR adjustments.
Adjust R14 and C4 for pulse response, and adjust C6 for low-frequency CMRR.
1 Change the function generator settings to:
•
•
•
Square wave
2.5 kHz
600 mV
p-p
44
Download from Www.Somanuals.com. All Manuals Search And Download.
Calibration Tests and Adjustment
Adjustment Procedure
2 Change the oscilloscope settings to:
Menu
Selection
(time/div)
(sensitivity)
Setting
TIMEBASE
CHAN 1
50 µs/div
100 mV/div
3 Carefully connect the input of the probe to the test board in the position shown
in the figure below (signal to + input).
Figure 2-17
Signal to + input
4 Adjust R14 (LF Gain) and C4 (+ LF BANDWIDTH) for the flattest pulse top
(see figure below). Again, adjust slowly and press CLEAR DISPLAY frequently
to restart averaging.
Figure 2-18
R14 and C4 Adjustment
45
Download from Www.Somanuals.com. All Manuals Search And Download.
Calibration Tests and Adjustment
Adjustment Procedure
5 Carefully connect the probe to the test board in position shown in the figure
below (signal to both inputs).
Figure 2-19
Signal to both inputs
6 Change the function generator settings to:
•
•
•
Sine wave
4 kHz
1.0 V
p-p
7 Change the oscilloscope settings to:
Menu
Selection
(time/div)
(sensitivity)
Setting
TIMEBASE
CHAN 1
50 µs/div
2 mV/div
8 Adjust C6 (-LF BANDWIDTH) for minimum signal amplitude on the
oscilloscope. Again, adjust C6 slowly and press CLEAR DISPLAY frequently to
restart averaging.
46
Download from Www.Somanuals.com. All Manuals Search And Download.
Calibration Tests and Adjustment
Adjustment Procedure
High Frequency Compensation
This adjustment sequence continues from the Low Frequency Response and CMRR
adjustment. However, it can be done separately if the probe meets all specifications
except bandwidth. Adjust R13 for unity gain at 200 MHz.
1 Connect the signal generator to the test board and set it for 200 MHz and
300mV (107 mV ).
p-p
rms
2 Carefully connect the input of the probe to the test board in the position shown
in the figure below (signal to + input).
Figure 2-20
Signal to + input
3 Press AUTOSCALE, then measure the peak-to-peak voltage on channel 1
(Press SHIFT (blue), press V P-P, then press 1).
4 Adjust R13 (HF COMP) to make the signal amplitude measurement on the
oscilloscope 300 mV , or as close to that as possible. Adjust slowly and press
p-p
CLEAR DISPLAY frequently to restart averaging.
Figure 2-21
R13 Adjustment
The minimum allowable amplitude is 212 mV . Typical values will be between 275 and
p-p
325 mV . the probe needs repair if the minimum cannot be reached.
p-p
Probe Reassembly
1 Disconnect the probe power cable at the rear panel of the 1142A
2 Remove the probe PC assembly from the test board. Be sure the probe input
connectors remain attached to the probe.
3 Remove the grounding screw and ground block from the PC assembly.
47
Download from Www.Somanuals.com. All Manuals Search And Download.
Calibration Tests and Adjustment
Adjustment Procedure
4 Assemble the PC assembly in the top cover.
The side of the PC assembly with the large hybrid is exposed when the assembly is in
the top cover. The figure below shows how the top cover, PC board, and ground block
fit together.
Figure 2-22
Reassembling the Probe
a
Insert the input connectors first, and seat the cable end of the PC assembly over the
pins at the rear of the cover.
b
c
Position the ground block at the center-front of the PC assembly.
Insert the grounding screw through the top cover and PC assembly and screw it into
the ground block as shown.
5 Replace the bottom cover.
Position the cable strain relief and with one hand, hold the cable and top together.
a
The flange on the strain relief has a notch that fits around a protrusion in the top cover.
C A U T I O N
Note where the two pins at the rear of the top cover enter the holes in the PC assembly.
Position the cable wires away from these two areas. Otherwise, when the bottom cover
is closed, part of it will pinch wires that are laying over these areas.
b
c
Insert the pins at the front of the bottom cover into the holes at the front of the top
cover.
Close the two covers together and fasten with the probe clamp ring.
48
Download from Www.Somanuals.com. All Manuals Search And Download.
Calibration Tests and Adjustment
Attenuator Adapter Adjustment
Attenuator Adapter Adjustment
The following procedure should be used if it is necessary to adjust an attenuator adapter.
Attenuator adapters have only characteristics; they do not have any specifications. An
adapter will need adjustment only if one or more of the following occurs.
•
If an adapter is to be used on a different 1141A probe that it was calibrated with
last.
•
•
If an adapter needs to be optimized to requirements for a special measurement.
If an adapter is suspected of needing adjustment.
N O T E
Attenuator adapters should not be adjusted as part of routine maintenance. The design
necessary to give the high CMRR and high impedance of the attenuators makes
adjustment delicate.
Additionally, because of broadband noise, a spectrum analyzer should be used to verify
adjustment quality.
Each attenuator contains two identical voltage dividers on a ceramic substrate. There
are three adjustments. The low-frequency CMRR adjust the low-frequency balance
between the two voltage dividers. The two high-frequency adjustments are electrically
identical. Each adjusts the high-frequency compensation of one of the voltage dividers.
To meet both pulse response and CMRR characteristics they are adjusted differently.
Briefly, the adjustment procedure is:
a
Adjustthepositive high-frequencyresponse(+HFRESP)forthe best pulse response
using a 3.5 kHz square wave.
b
Adjust the negative high-frequency response (-HF RESP) and Low-frequency CMRR
(LF CMRR) for best CMRR using a 3.5 kHz square wave.
The following equipment is required for this procedure. Procedures are based on the
model or part number recommended.
Equipment Required
Equipment Required
Critical Specifications
3.5 kHz, 3 V to 16 V
Recommended
Model/Part
Function Generator
Oscilloscope
Probe/Power Module
Test Board
33120A
p-p
100 MHz at 1 mV/div
No substitute
54830B
1141A/1142A
01141-66504
10502A
No substitute
Cable
BNC, 50 Ω 9-inch
BNC, 50 Ω 36-inch
Cable
10503A
Alignment tool
Flat blade (supplied accessory) 8710-1961/
Sprague-Goodman
part, GTT-5G
49
Download from Www.Somanuals.com. All Manuals Search And Download.
Calibration Tests and Adjustment
Attenuator Adapter Adjustment
Adjustment Procedure
N O T E
The attenuator must be adjusted when installed on the 1141A probe with which it will
be used. The specifications and characteristics will not be met if the attenuator adapter
is adjusted with one differential probe then used with another.
1 Removetheprobepinsfromtheattenuatoradapteranddifferentialprobe, then
attach the adapter to the probe.
2 Set the 1142A front panel switches to Local and Zero offset.
3 Use the 9-inch BNC cable to connect the function generator to the test board.
The short cable minimized ground-loop voltages.
4 Set up the function generator.
•
•
•
Square wave
3.5 kHz
3.0 Vp-p for 10x adapter and 16 V for a 100x adapter.
p-p
5 Use the long BNC cable to connect the Trig Out of the function generator to
the EXT TRIG of the oscilloscope.
6 Set up the oscilloscope, then set the following parameters.
Menu
Selection
Setting
TIMEBASE
CHAN 1
(time/div)
50 µs/div
(sensitivity)
(input R)
50 mV/div
50 Ω DC
TRIG
(mode)
source
level
trg’d
EXT
1.00000 V
ACQUISITION
Sampling Mode
Memory Depth
Sample Rate
Averaging
Real Time
Automatic
Automatic
Enabled
32
# of avg
50
Download from Www.Somanuals.com. All Manuals Search And Download.
Calibration Tests and Adjustment
Attenuator Adapter Adjustment
7 Connect the adapter/probe combination to the test boardinthe positionshown
in the figure below.
Figure 2-23
Signal to + input
8 Adjust the + HF RESP for best overall pulse response, the flattest pulse top.
Use the figure below for adjust locations.
Figure 2-24
Adjustment Locations
9 Change the function generator to 10 V (10x adapter adjustment only).
p-p
10 On the oscilloscope, press CHAN and set the sensitivity to 1 mV/div.
51
Download from Www.Somanuals.com. All Manuals Search And Download.
Calibration Tests and Adjustment
Attenuator Adapter Adjustment
11 Connect the adapter/probe combination to the test boardinthe positionshown
in the figure below (signal to both inputs).
Figure 2-25
Signal to both input
12 Alternately adjust the LF CMRR and then the -HF RESP for a minimum signal
on the oscilloscope. Repeat the adjustments until the signal is optimized to a
minimum. Each adjustment should be set to minimize the component of the
signal it affects most. Some high-frequency components of the signal are not
affected by either adjustment.
52
Download from Www.Somanuals.com. All Manuals Search And Download.
Service
Introduction
Introduction
This section provides troubleshooting, service, and repair information for the
1141A Differential Probe and 1142A Probe Control and Power Module. The
troubleshooting information is provided to isolate a faulty assembly. When a
faulty assembly has been located, the disassembly/assembly procedures help
direct replacement of the assembly.
W A R N I N G
Maintenance should be performed by trained service personnel aware of the hazards
involved (for example, fire and electric shock). When maintenance can be performed
without power applied, the power cord must be removed from the instrument.
54
Download from Www.Somanuals.com. All Manuals Search And Download.
Service
Performance Specifications and Characteristics
Performance Specifications and Characteristics
The following table gives performance specifications used to test the 1141A and 1142A.
It also gives performance characteristics that are typical for the probe system.
Performance Specifications and Characteristics
Parameter
Probe alone
With 10x attenuator
With 100x attenuator
SPECIFICATIONS
1
Bandwidth (-3 dB, dc coupled) dc to 200 MHz
Rise Time: (calculated)
Gain Accuracy
CMRR
1.75 ns
2.0%
4.0%
500 V(dc + peak ac)
See the graphs in figure 3-1
200 V(dc + peak ac)
Maximum Input Voltage
(see figure 3-2)
Differential Input Range
DC mode with no offset
300 mV peak
3.0 V peak
30 V peak
with DC Reject or appropriate 20 Vdc, decreasing to 200 Vdc, decreasing to 500 Vdc, decreasing to
offset
300 mV at 30 Hz
3.0 mV at 30 Hz
30 mV at 30 Hz
Common-mode Operating
Range
dc
20 Vdc
linear change
0.5 V
200 Vdc
linear change
5 V
500 Vdc
linear change
50 V
dc to 30 Hz
30 Hz to 200 MHz
dc Offset Range
20 V
200 V
500 V
Input Impedance Resistance
Capacitance
1 MΩ
7 pF
9 MΩ
3.5 pF
10 MΩ
2 pF
ac Low-freq. Response (-3dB)
dc Reject Response
Output Impedance
Thermal drift
15 Hz
1.5 Hz
1.5 Hz
5 Hz, 0.5 Hz, or 0.05 Hz (selectable irrespective of attenuator)
50 Ω
≤ 50 µVdc/°C
Displayed noises
≤ 50 µV
rms
Overload Recovery
< 1 ms from overdrive that is less than the common mode range
Note: 1. For maximum signal fidelity above 100 MHz, limit the probe input (without attenuators to ≤ 300 mV
peak-to-peak.
55
Download from Www.Somanuals.com. All Manuals Search And Download.
Service
Performance Specifications and Characteristics
Figure 3-1
CMRR Specifications and Characteristics
Legend
A.
B.
C.
D.
E.
CMRR specification for probe with no input adapters.
Low-frequency CMRR specification for probe with the ac adapter.
Typical CMRR characteristic for differential probe with no input adapters
Typical CMRR characteristic for differential probe with 100x attenuator adapter at input.
Typical CMRR characteristic for differential probe with 100x attenuator adapter at input.
56
Download from Www.Somanuals.com. All Manuals Search And Download.
Service
Performance Specifications and Characteristics
Figure 3-2
Maximum Input Voltage vs. Frequency
Legend
A.
B.
C.
Input voltage limits for probe alone.
Input voltage limits for 10x adapter.
Input voltage limits for 100x adapter.
57
Download from Www.Somanuals.com. All Manuals Search And Download.
Service
General Characteristics
General Characteristics
Thefollowingcharacteristics apply tothe 1141A Differential Probewith the 1142AProbe
Control and Power Module.
Environmental Conditions
Operating
Non-operating
Temperature
Humidity
0°C to +55 C° (32°F to +131°F)
-40°C to +70°C (-40°F to +158°F)
up to 95% relative humidity (non-
condensing) at +40°C (+104°F)
up to 90% relative humidity at +65°C
(+149°F)
Altitude
up to 4,600 meters (15,000 ft)
up to 15,300 meters (50,000 ft)
Vibration
Random vibration 5 to 500 Hz, 10
minutes per axis, 0.3grms.
Random vibration 5 to 500 Hz, 10 min.
per axis, 2.41 grams. Resonant search
5 to 500 Hz swept sine, 1Octave/min.
sweep rate, (0.75g), 5 min. resonant
dwell at 4 resonances per axis.
Power
Requirements
Voltage: 90 to 132/198 to 264 Vac, 47 to 440 Hz
Power: 25 VA maximum
Weight
Net: approximately 1.8 kg (4.0 lb.)
Shipping: approximately 2.7 kg (6.0 lb.)
Dimensions
Refer to the outline drawings below.
Figure 3-3
1142A Probe Control
and Power Module
1141A Differential
Probe
Mechanical Dimensions
58
Download from Www.Somanuals.com. All Manuals Search And Download.
Service
Theory of Operation
Theory of Operation
The following discussion covers block-level theory for the 1141A/1142A differential
probe system. Refer to the block diagram below.
The differential probe system consists of two units, the 1141A Differential Probe with
its accessories and the 1142A Probe Control and Power Module. For purposes of the
following discussion, these will be called the probe and the control module respectively.
Figure 3-4
Differential Probe System Block Diagram
59
Download from Www.Somanuals.com. All Manuals Search And Download.
Service
Theory of Operation
Differential Probe
The probe contains a two-path differential amplifier with unity gain. It is implemented
on a double-sided surface-mount PC board with the high-frequency path on one side and
the low-frequency path on the other. The two paths are split directly after the differential
input connections.
High-Frequency Path
The positive and negative inputs are ac-coupled at 33 Hz into identical impedance
converters. The HF CMRR adjustment balances the gain at the outputs of the impedance
converters. The impedance feed a semi-custom differential amplifier. An additional
negative input to the differential amplifier brings in the sum of the low-frequency and
feedbacksignals.TheFREQCOMPadjustmentprovidesvariablehigh-frequencypeaking
of the differential amp. The output amp provides two signals. A feedback signal is
summed with the low-frequency signals and the output signal is the final output of the
probe. Overall probe gain is set by the HIGH FREQ GAIN adjustment.
The entire signal portion of the high-frequency path is implemented on a hybrid IC.
Support circuitry includes bias for the impedance converters and a bias supply for
current sources on the hybrid.
Low-Frequency Path
The bandwidth of the low-frequency path is approximately 75 KHz. The positive and
negative inputs are dc coupled into identical inverting op-amps with gain of 0.5. They
provide a precision 1 M Ω input impedance for the probe. The LOW FREQ BANDWIDTH
adjustments match the gain and phase of the low-frequency path to that of the feedback
from the probe output. One of the two adjustments is set to match the properties of the
feedback and the other is to match the two low-frequency paths. These adjustments
affect the CMRR quality of the probe. The inputs of the inverting op-amps include
protection for ESD and over-voltage conditions.
The inverting op-amps feed a precision differential amplifier with unity gain and a single-
ended output. The output is fed to the summing amp and to the control module for use
in the dc reject circuit.
Summing Amp
The summing amp combines the feedbacksignal, the low-frequencysignal, andthe offset
signal. The LOW FREQ GAIN adjustment matches the gain of the low-frequency path to
the overall gain.
Control and Power Module
The control and power module provides offset functions, local and remote control, and
power to the probe system.
60
Download from Www.Somanuals.com. All Manuals Search And Download.
Service
Theory of Operation
Offset Functions
There are two offset functions developed in the control module: variable offset and dc
reject. A variable offset voltage with coarse and fine adjustments can be selected by the
front panel controls. The offset level is buffered by U8 and selected by multiplexer U3
as the input to offset amp U7. The output of the offset amp is summed with the low-
frequency signal and feedback which gives dc coupling in the probe. Front panel
screwdriver adjustment Offset Null zeros the dc output from the probe when the dc input
and offset are zero.
For dc Reject, an output from the low-frequency amplifier in the probe (LFSIG) is used
to develop a voltage used to null the dc component of the input signal. LFSIG is an input
to U6, an inverting amplifier and low-pass filter. Multiplexer U3 selects one of three
capacitors to set a roll-off frequency of 0.05, 0.5, or 5 Hz. The output of U6 is selected,
again by U3, as the input to the offset amp. When the output of the offset amp is summed
into the low frequency path, the result is cancellation of the dc component of the input
signal. Front panel screwdriver adjustment DC Reject Gain adjusts the gain of the dc
reject circuit.
Local and Remote Control
The front panel switch controls the dc reject and offset functions. It also selects remote
operation, which allows control through the rear panel remote input connector.
Power Supply
The supply provides 6 V and 15 V for the probe and analog control circuitry as well
as +5 V for the digital control circuitry.
Attenuator Adapters
The 10x and 100x Attenuator Adapter are similar. A ceramic substrate carries two
attenuators, one for each input polarity. A variable resistor adjusts the low-frequency
balance (LF CMRR) between the two attenuators. The high-frequency adjustments are
the same for each attenuator. Each attenuator is adjusted differently. One attenuator is
adjusted for optimum pulse response and the other for best high-frequency CMRR.
Figure 3-5
Attenuator Adapter, Simplified Schematic
61
Download from Www.Somanuals.com. All Manuals Search And Download.
Service
Theory of Operation
Test Board
The test board is a device for conveniently connecting test signals to the differential
probe. The probe can be connected to the board with the signal to the positive, negative,
or both inputs.
Figure 3-6
Test Board Schematic
62
Download from Www.Somanuals.com. All Manuals Search And Download.
Service
Service Policy
Service Policy
For parts of the 1141A/1142A probe system that are complex, the service policy is for
assembly-level repair. For parts of the system with simple circuitry, the service policy is
component-level repair.
The service policy for the 1141A Differential Probe is assembly-level repair. Assemblies
include the PC assembly and cable. The PC assembly is an “exchange assembly.” A
repaired and tested assembly is shipped upon receipt of the defective assembly.
The attenuator and ac coupling adapters are shipped as complete assemblies.
The service policy for the 1142A Probe Control and Power Module is component-level
repair.
63
Download from Www.Somanuals.com. All Manuals Search And Download.
Service
Troubleshooting
Troubleshooting
Usethefollowingparagraphstoassistintroubleshootingproblemswiththe1141A/1142A
Differential Probe.
Probe Troubleshooting
To troubleshoot the probe:
1 Apply a known signal to the input of the probe.
2 Check for an identical output at the output coax to the cable. This connection
is the one soldered to the PC board. If the probe output cable is not terminated,
or the coax is open, the output signal will be about twice the amplitude of the
input signal.
3 Ifthesignalisincorrect, checkthepowersupplyvoltagesfromthe1142AProbe
Control and Power Module. Use the cable diagram on the next page.
4 Troubleshoot the cable with an ohmmeter. Use the cable diagram on the next
page.
Probe Control and Power Module
Troubleshooting
The circuitry consists of simple power supplies, operational amplifiers, and TTL. Use
conventional troubleshooting techniques. A complete parts list, component locator, and
schematics are provided later in this chapter.
64
Download from Www.Somanuals.com. All Manuals Search And Download.
Service
Removing and Replacing Assemblies
Removing and Replacing Assemblies
This section contains procedures for the removal and replacement of major assemblies.
C A U T I O N
Never remove or install any assembly with the instrument power ON. Component
damage can occur.
Differential Probe
Use the following procedure to remove and replace the amplifier PC board in the
differential probe.
C A U T I O N
C A U T I O N
ELECTROSTATIC DISCHARGE can damage electronic components. Use grounded
wrist straps and mats when servicing the probe.
Handle the differential probe carefully once it has been disassembled. If unsupported,
the weight of the cable can put strain on the PC board.
Disassemble Probe
1 Remove the probe tip caps and probe tips.
2 Loosen the probe clamp ring at the cable end of the probe (1/4 turn counter-
clockwise) and slide it down the cable.
3 Remove the bottom cover.
a
At the cable end of the probe, separate the covers about one centimeter (1/2 inch).
b
Slide the bottom cover toward the cable end of the probe until the locator pins at
the probe input clear the holes. Then, remove the cover.
Note the way the cable strain relief is keyed and held at the rear of the top cover.
4 Remove the probe top cover.
The ground screw passes through the top cover and PC board and screws into the ground
block.
a
Remove the ground screw on the top of the probe.
At the cable end, the PC board fits over the pins in the top cover.
b
Lift the board off of the pins and slide it in the direction of the cable until the input
connectors clear the front of the probe.
5 Un-solder the two connections where the coaxial output cable connects to the
PC board.
6 Disconnect the cable connector from the probe PC board.
66
Download from Www.Somanuals.com. All Manuals Search And Download.
Service
Removing and Replacing Assemblies
Reassemble Probe
The ground screw passes through the top cover and PC board and screws into the ground
block.
1 If replacing the PC board, remove the input connectors from the old board and
put them on the new one.
2 If replacing the cable, note the orientation of the probe clamp ring on the old
cable, remove the ring and put it on the new cable.
3 Connect the cable connector to the PC board.
4 Solder the two connections of the coaxial cable to the PC board.
On a new probe cable, the conductors of the coaxial cable are connected by a heavy
single wire. Cut the heavy wire so it matches the wire on the cable that was removed.
5 Assemble the PC assembly into the top cover.
The large hybrid is exposed when the assembly is in the top cover. The figure on the
below shows the sequence of the ground screw, top cover, PC board, and ground block.
Figure 3-8
Reassembling the Probe
a
Insert the input connectors first, and seat the cable end of the PC assembly over the
pins at the rear of the cover.
b
c
Position the ground block at the front of the PC assembly.
Insert the grounding screw through the top cover and screw it into the ground block
as shown in the figure above.
6 Replace the bottom cover.
a
Position the cable strain relief and with one hand, hold the cable and top cover
together.
The flange on the strain relief has a notch that fits over a protrusion in the top cover.
67
Download from Www.Somanuals.com. All Manuals Search And Download.
Service
Removing and Replacing Assemblies
C A U T I O N
Note where the two pins at the rear of the top cover enter the holes in the PC assembly.
Position the cable wires away from these two areas. When the bottom cover is closed,
part of it will pinch wires that are laying over these areas.
b
c
Insert the pins at the front of the bottom cover into the holes at the front of the top
cover.
Close the covers together and fasten with the probe clamp ring. If the covers do not
fit together tightly and easily, check for pinched wires (see caution above).
Probe Adapters
Use the following procedure to disassemble the probe adapters. The adapter housing
consists of two plastic parts, one of which slides into the other. The parts are held
together by the spring effect of two plastic tabs on the inner part.
Mechanically, the ac adapter is about the same as the 10x and 100x attenuator adapters.
The attenuator adapters have an extra ground connector which connects the substrate
ground to the thumbwheel screw and plating inside the housing.
Disassemble Adapter
1 Remove the probe tip caps and probe tips from the adapter input.
2 Hold the adapter in one hand taking care not to block the output end of the
adapter (the end which attaches to the probe).
3 Note the view of the input end of the adapter in the figure below. The arrows
indicate the holding tabs.
Figure 3-9
Disassembling Adapters
4 With the thumb and forefinger, squeeze the tabs together, as indicated by the
arrows. Simultaneously, push the tabs into the outer housing so the inner
housing begins to slide out.
5 While holding the outer housing, push back against the thumbwheel until the
inner housing can be grasped and removed.
Reassemble Adapter
Reassembling the adapter is slightly harder because you have to align the connector pins
and thumbwheel screw, while sliding the inner housing and outer housing together.
1 Be sure the input connectors and output pins are present and seated on the
substrate or PC board.
68
Download from Www.Somanuals.com. All Manuals Search And Download.
Service
Removing and Replacing Assemblies
2 Combine the thumbwheel screw and thumbwheel and insert them into the hole
in the outer housing.
3 Insert the substrate/board into the outer housing. Slip the attenuator ground
(attenuator adapters only) over the thumbwheel screw and seat the input
connectors in the proper holes in the housing.
Figure 3-10
Reassembling the Adapter
Do not force reassembly of the adapter. The housing halves will slide together with
moderate friction.
4 Align the inner housing tabs with the grooves in the side of the outer housing
and slide the two partly together.
While seating the thumbwheel screw and pins, it will help to hold the assembly vertically,
with the input end of the adapter down. This will allow the assemblies to sit vertically
and more easily align with the holes in the inner housing.
5 Seat the thumbwheel screw first, then the output pins, into the appropriate
holes in the inner housing.
6 Once the two housings are nearly together, press them together firmly until
the tabs click into place.
69
Download from Www.Somanuals.com. All Manuals Search And Download.
Service
Removing and Replacing Assemblies
Probe Control and Power Module
Use the following procedure to disassemble the probe control and power module.
W A R N I N G
W A R N I N G
Hazardous voltages exist on the power supply. To avoid electrical shock, adhere closely
to the following procedures.
1 Remove the power cord.
2 Remove four flathead screws and remove the top cover.
Be sure to reconnect the safety ground when reassembling the instrument.
3 Unplug the safety ground from the tab on the rear panel of the instrument.
4 Note the orientation of the knobs. Remove the two knobs.
5 On the bottom of the instrument, remove the 5mm screw that fastens the
transformer support.
6 Remove the following pan-head screws.
•
•
•
Three directly on the PC board.
Two on the ac input connector.
Two on the heatsink.
7 Remove the PC board. Slide it slightly forward so parts will clear the rear panel,
then lift the rear of the board out while sliding it backwards.
8 Remove the two heatsink spacers from the standoffs that were directly under
the heatsink.
9 Reverse the procedure to reassemble the control and power supply.
70
Download from Www.Somanuals.com. All Manuals Search And Download.
Service
Replaceable Parts
Replaceable Parts
This section contains information for ordering parts. Service support for the 1141A
DifferentialProbeistotheassemblylevel. Servicesupportfor the adaptersisas complete
assemblies, except for the probe tip caps and the probe tips. Service support for the
1142A Probe Control and Power Module is to the component level.
Parts List
The replaceable parts lists include all parts relevant to the applicable service levels. The
information given for each part consists of the following:
•
•
•
Reference designator
Part number
Total quantity (Qty) in instrument or on assembly. The total quantity is given once
and at the first appearance of the part number in the list.
•
•
Description of part
Typical manufacturer of part in a five-digit code.
Ordering Information
To order a part in the material part list, quote the part number, indicate the quantity
desired, and address the order to the nearest Agilent Technologies Sales Office.
To order a part not listed in the material list, include the instrument part number,
instrument serial number, a description of the part (including its function), and the
number of parts required. Address the order to the nearest Agilent Technologies Sales
Office.
Direct Mail Order System
Within the USA, Agilent Technologies can supply parts through a direct mail order
system. There are several advantages to this system:
•
Direct ordering and shipment from the Agilent Technologies parts center in
California, USA.
•
No maximum or minimum on any mail order (there is a minimum amount for parts
ordered through a local Agilent Technologies Sale Office when the orders require
billing and invoicing).
•
•
Prepaid transportation (there is a small handling charge for each order).
No invoices.
In order for Agilent Technologies to provide these advantages, a check or money order
must accompany each order.
71
Download from Www.Somanuals.com. All Manuals Search And Download.
Service
Replaceable Parts
Mail order forms and specific ordering information are available through your local
Agilent Technologies Sales Office. Addresses and telephone numbers are located in a
separate document shipped with the manuals.
Manufacturers’ Codes
A list of manufacturers’ codes is given the table below. The codes are given for parts in
the parts lists. The table gives the manufacturer and address for each code.
Manufacturers’ Code List
Mfr. No.
Name
Address
00000
06665
24546
27014
28480
32997
Any satisfactory supplier
Precision Monolithics Inc.
Corning Glass Works (Bradford)
National Semiconductor Corp
Agilent Technologies Corporate Hq
Bourns Inc.
Santa Clara, CA 95050
Bradford, PA 16701
Palo Alto, CA 94304
Palo Alto, CA 94304
Riverside, CA 92507
72
Download from Www.Somanuals.com. All Manuals Search And Download.
Service
Replaceable Parts
1141A and 1142A Replaceable Parts
Ref.
Des.
Part
Number
Qty
Description
Mfr.
Code
Mfr. Part
Number
1141A Differential Probe
A1
01141-66503
01141-26102
01141-22301
01141-22401
01141-24102
01141-44702
01141-44102
01141-22502
01141-94307
1
2
1
1
2
1
1
1
1
PC ASSEMBLY-DIFFERENTIAL PROBE
PROBING PIN
28480 01141-66501
28480 01141-26102
28480 01141-22301
28480 01141-22401
28480 01141-24101
28480 01141-44701
28480 01141-44101
28480 01141-22501
28480 01141-94301
E1
E2
GROUND BLACK
E3
GROUND SCREW
MP1
MP2
MP3
MP4
MP5
CAP-PROBING PIN
PROBE BOTTOM COVER
PROBE TOP COVER
PROBE POS CLAMP RING
DIFFERENTIAL PROBE LABEL
W1
01141-61603
1
PROBE CABLE ASSEMBLY
28480 01141-61601
10X Attenuator Adapter
PROBING PIN
01141-26102
01141-24102
2
2
28480 01141-26102
28480 01141-24101
CAP-PROBING PIN
100X Attenuator Adapter
PROBING PIN
01141-26102
01141-24102
2
2
28480 01141-26102
28480 01141-24101
CAP-PROBING PIN
AC Adapter
01141-26102
01141-24102
2
2
PROBING PIN
CAP-PROBING PIN
28480 01141-26102
28480 01141-24101
1142A PROBE CONTROL AND POWER MODULE
78
Download from Www.Somanuals.com. All Manuals Search And Download.
Service
Replaceable Parts
1141A and 1142A Replaceable Parts
Ref.
Des.
Part
Number
Qty
Description
Mfr.
Code
Mfr. Part
Number
A1
01142-66501
0515-0374
0515-1031
0515-1579
1
7
4
1
PC ASSEMBLY-POWER AND CONTROL
SCREW-MACHINE M3 10mm-LG
28480 01142-66501
H1
H2
H3
00000 ORDER BY DESP.
00000 ORDER BY DESP.
00000 ORDER BY DESP.
SCREW-MACHINE M3 6mm-LG 90-DEG-FLH-HD
SCREW-MACHINE M5 18mm-LG
MP1
MP2
MP3
MP4
MP5
MP6
01142-47702
01142-44101
01142-24701
0370-1097
1
1
2
2
4
5
BOTTOM CHASSIS
TOP CHASSIS
SPACER-HEAT SINK
KNOB-POINTER
FOOT
28480 01142-47701
28480 01142-44101
28480 01142-24701
28480 0340-1097
28480 0403-0727
28480 5041-0234
0403-1012
5041-0234
KEYCAP
W1
8120-1521
1
POWER CORD 18-AWG 3-COND 90-IN-LG (US/
Canada only)
28480 8120-1521
79
Download from Www.Somanuals.com. All Manuals Search And Download.
Service
Replaceable Parts
Power Supply Replaceable Parts
Ref. Des.
Part
Number
Qty
Description
Mfr.
Code
Mfr. Part
Number
Prefix the reference designators with A1
CAPACITOR-FXD 0.33UF 10% 50VDC
CAPACITOR-FXD 0.1UF 5% 100VDC MET-POLY
CAPACITOR-FXD 2200UF+30-10% 50VDC AL
CAPACITOR-FXD 2200UF+30-10% 50VDC AL
CAPACITOR-FXD 0.01UF 10% 100VDC CER
CAPACITOR-FXD 0.01UF 10% 100VDC CER
CAPACITOR-FXD 0.01UF 10% 100VDC CER
CAPACITOR-FXD 0.01UF 10% 100VDC CER
CAPACITOR-FXD 4.7UF 10% 35VDC TA
CAPACITOR-FXD 4.7UF 10% 35VDC TA
NOT ASSIGNED
C1
0160-6190
0160-5474
0180-3298
0180-3298
0160-6500
0160-6500
0160-6500
0160-6500
0180-3845
0180-3845
2
1
2
28480 0160-6190
28480 0160-5474
28480 0180-3298
28480 0180-3298
28480 0160-6500
28480 0160-6500
28480 0160-6500
28480 0160-6500
28480 0180-3845
28480 0180-3845
C2
C3
C4
C5
4
C6
C7
C8
C9
4
1
C10
C11
C12
C13
C14
C15
C16
C17
C18
C19
C20
C21-22
C23
C24
C25
C26
0160-5581
0160-6190
0160-7060
0160-4801
0180-3845
0180-3845
CAPACITOR-FXD 0.033UF 10% 63VDC
CAPACITOR-FXD 0.33UF 10% 50VDC
CAPACITOR-FXD 3.3UF 10% 63VDC
CAPACITOR-FXD 100PF 5% 100VDC CER
CAPACITOR-FXD 4.7UF 10% 35VDC TA
CAPACITOR-FXD 3.3UF 10% 63VDC
NOT ASSIGNED
28480 0160-5581
28480 0160-6190
28480 0160-7060
28480 0160-4801
28480 0180-3845
28480 0180-3845
1
1
0160-5471
0160-5469
1
1
CAPACITOR-FXD 0.1UF 5% 50VDC MET-POLY
CAPACITOR-FXD 1UF 10% 50VDC MET-POLY
NOT ASSIGNED
28480 0160-5471
28480 0160-5469
0180-3784
0180-3784
0180-3784
0180-3784
4
CAPACITOR-FXD 22UF 20% 25VDC TA
CAPACITOR-FXD 22UF 20% 25VDC TA
CAPACITOR-FXD 22UF 20% 25VDC TA
CAPACITOR-FXD 22UF 20% 25VDC TA
28480 0180-3784
28480 0180-3784
28480 0180-3784
28480 0180-3784
80
Download from Www.Somanuals.com. All Manuals Search And Download.
Service
Replaceable Parts
Power Supply Replaceable Parts
Ref. Des.
Part
Number
Qty
Description
Mfr.
Code
Mfr. Part
Number
CR1
NOT ASSIGNED
CR2
1901-1087
1901-1087
2
DIODE-PWR RECT 100V 3A
DIODE-PWR RECT 100V 3A
NOT ASSIGNED
04713 MR501
04713 MR501
CR3
CR4
CR5
1901-0731
1901-0731
8
8
DIODE-PWR RECT 400V 1A
DIODE-PWR RECT 400V 1A
NOT ASSIGNED
28480 1901-0731
28480 1901-0731
CR6
CR7-8
CR9
1901-0731
1901-0731
1901-0731
1901-0731
1901-0734
1901-0734
8
8
8
8
2
2
DIODE-PWR RECT 400V 1A
DIODE-PWR RECT 400V 1A
DIODE-PWR RECT 400V 1A
DIODE-PWR RECT 400V 1A
DIODE-PWR RECT 1N5818 30V 1A
DIODE-PWR RECT 1N5818 30V 1A
NOT ASSIGNED
28480 1901-0731
28480 1901-0731
28480 1901-0731
28480 1901-0731
04713 IN5818
CR10
CR11
CR12
CR13
CR14
CR15-16
CR17
CR18
04713 IN5818
1901-0731
1901-0731
DIODE-PWR RECT 400V 1A
DIODE-PWR RECT 400V 1A
28480 1901-0731
28480 1901-0731
DS1
1990-0521
1
LED-LAMP LUM-INT=2.2MCD IF=50MA-MAX
24840 5082-4955
E1
E2
2110-0642
2110-0565
1
1
FUSEHOLDER 6.3A 250A
FUSEHOLDER CAP
28480 2110-0642
28480 2110-0565
F1
2110-0201
0515-1579
01142-21101
1
1
1
FUSE 0.25A 250V TD
SCREW-MACHINE M5 18mm-LG
HEAT SINK
28480 2110-0201
28480 0515-1579
28480 01142-21101
H1
HS1
81
Download from Www.Somanuals.com. All Manuals Search And Download.
Service
Replaceable Parts
Power Supply Replaceable Parts
Ref. Des.
Part
Number
Qty
Description
Mfr.
Code
Mfr. Part
Number
J1
J2
J3
1252-4731
1252-1487
1252-3935
1
1
1
CONNECTOR-AC PWR
28480 1251-4743
28480 1252-1487
28480 1252-3134
CONN-RECT D-SUBMIN 9-CKT (remote)
CONNECTOR-ROUND 8-CKT (probe)
MP1
MP2
MP3
MP4
MP5-6
MP7
MP8
MP9
1400-1604
1205-0732
0361-0685
0340-1211
1
4
3
3
LED MOUNT
28480 1400-1604
28480 1205-0732
28480 0361-0685
28480 0340-1211
SPRING CLIP
RIVET-BLIND DR-PIN RNDH 0.125DIA
INSULATOR-THERMAL
NOT ASSIGNED
1400-0249
1
1
1
CABLE TIE 0.062-0.625-DIA 0.091-WD NYL
TRANSFORMER SUPPORT
WASHER-TRANSFORMER SUPPORT
16956 08-465/GRY
28480 01142-24702
28480 01142-28801
01142-24702
01142-28801
R1-2
R3
NOT ASSIGNED
0757-0442
0757-0465
0757-0199
0757-0199
0698-4431
0699-1203
0757-0434
0698-4431
0699-1203
0757-0434
0757-0420
0698-4002
0699-1203
4
1
5
RESISTOR 10K 1% 0.125W TF TC=0 100
RESISTOR 100K 1% 0.125W TF TC=0 100
RESISTOR 21.5K 1% 0.125W TC=0 100
RESISTOR 21.5K 1% 0.125W TC=0 100
RESISTOR 2.05K 1% 0.125W TF TC=0 100
RESISTOR 120.0 1% 0.125W TF TC=0 25
RESISTOR 3.65K 1% 0.125W TF TC= 100
RESISTOR 2.05K 1% 0.125W TF TC=0 100
RESISTOR 120.0 1% 0.125W TF TC=0 25
RESISTOR 3.65K 1% 0.125W TF TC=0 100
RESISTOR 750 1% 0.125W TF TC=0 100
RESISTOR 5K 1% 0.125W TF TC=0 100
RESISTOR 120.0 1% 0.125W TF TC=0 25
24546 CT4-1/8-TO-1002-F
24546 CT4-1/8-TO-1003-F
24546 CT4-1/8-TO-2152-F
24546 CT4-1/8-TO-2152-F
24546 CT4-1/8-TO-2052-F
28480 0699-1203
R4
R5
R6
R7
2
4
2
R8
R9
24546 CT4-1/8-TO-3651-F
24546 CT4-1/8-TO-2052-F
28480 0699-1203
R10
R11
R12
R13
R14
R15
24546 CT4-1/8-TO-3651-F
24546 CT4-1/8-TO-751-F
24546 CT4-1/8-TO-5001-F
28480 0699-1203
1
2
82
Download from Www.Somanuals.com. All Manuals Search And Download.
Service
Replaceable Parts
Power Supply Replaceable Parts
Ref. Des.
Part
Number
Qty
Description
Mfr.
Code
Mfr. Part
Number
R16
R17
R18
R19
R20
R21
R22
R23
R24
R25
R26
R27
R28
R29
R30
R30
R32
R33-39
R40
R41
R42
R43
R44
R45
R46
R47
R48
0698-6317
0698-4002
0699-1203
0698-6317
2
RESISTOR 500 0.1% 0.125W TF TC=0 25
RESISTOR 5K 1% 0.125W TF TC=0 100
RESISTOR 120.0 1% 0.125W TF TC=0 25
RESISTOR 500 0.1% 0.125W TF TC=0 25
NOT ASSIGNED
28480 0698-6317
24546 CT4-1/8-TO-5001-F
28480 0699-1203
28480 0698-6317
0698-8827
0757-0442
0757-0427
0757-0401
2100-3161
2100-3056
0757-0199
0757-0199
0757-0442
0757-0442
0757-0280
0757-0460
1
RESISTOR 1M 1% 0.125W TF TC=0 100
RESISTOR 10K 1% 0.125W TF TC=0 100
RESISTOR 1.5K 1% 0.125W TF TC=0 100
RESISTOR 100 1% 0.125W TF TC=0 100
RESISTOR-TRMR 20K 10% TKF SIDE-ADJ
RESISTOR-TRMR 5K 10% TKF SIDE-ADJ 17-TRN
RESISTOR 21.5K 1% 0.125W TF TC=0 100
RESISTOR 21.5K 1% 0.125W TF TC=0 100
RESISTOR 10K 1% 0.125W TF TC=0 100
RESISTOR 10K 1% 0.125W TF TC=0 100
RESISTOR 1K 1% 0.125W TF TC=0 100
RESISTOR 61.9K 1% 0.125W TF TC=0 100
NOT ASSIGNED
28480 0698-8827
24546 CT4-1/8-TO-1002-F
24546 CT4-1/8-TO-1501-F
24546 CT4-1/8-TO-101-F
32997 3006P-1-203
1
1
1
1
32997 3006P-1-502
24546 CT4-1/8-TO-2152-F
24546 CT4-1/8-TO-2152-F
24546 CT4-1/8-TO-1002-F
24546 CT4-1/8-TO-1002-F
24546 CT4-1/8-TO-1001-F
24546 CT4-1/8-TO-6192-F
1
1
0757-0123
2100-4250
2100-4250
0757-0458
0683-2755
0698-4517
0698-3271
0698-8961
0757-0199
1
2
RESISTOR 34.8K 1% 0.125W TF TC=0 100
RESISTOR-VAR 10K 20%
28480 0757-0123
28480 2100-4250
RESISTOR-VAR 10K 20%
28480 2100-4250
1
1
1
1
1
RESISTOR 51.1K 1% 0.125W TF TC=0 100
RESISTOR 2.7M 1% 0.25W TF TC=0 100
RESISTOR 127K 1% 0.125W TF TC=0 100
RESISTOR 115K 1% 0.125W TF TC=0 100
RESISTOR 909K 1% 0.125W TF TC=0 100
RESISTOR 21.5K 1% 0.125W TF TC=0 100
24546 CT4-1/8-TO-5112-F
28480 0683-2755
24546 CT4-1/8-TO-1273-F
24546 CT4-1/8-TO-1153-F
28480 0698-8961
24546 CT4-1/8-TO-2152-F
83
Download from Www.Somanuals.com. All Manuals Search And Download.
Service
Replaceable Parts
Power Supply Replaceable Parts
Ref. Des.
Part
Number
Qty
Description
Mfr.
Code
Mfr. Part
Number
RP1
1810-1242
1
RESISTOR NETWORK
28480 1810-1242
S1
S2
3101-2609
3101-3007
1
1
SWITCH-SL DPST STD 5A 250VAC PC
SWITCH-6 STATION ASSEMBLY
28480 3101-2609
28480 3101-3007
T1
9100-4750
1
TRANSFORMER-POWER (with mtg. hardware)
28480 9100-4750
U1
U2
U3
U4
U5
U6
U7
U8
U9
U10
U1
1826-1403
1826-1670
5081-9233
1826-1403
1826-1670
1826-1381
1826-1381
1826-1381
1820-3177
1826-0774
1826-0635
2
2
1
IC V RGLTER-ADJ-POS 3/40V
00000 LT317AT
IC V RGLTR-ADJ-NEG -37/1.2V TO-220 PKG
ANALOG MULTIPLEXER 4 CHNL 16 -DIP-P
IC V RGLTR ADJ-POS 3/40V
00000 LT337AT
28480 1820-2182
00000 LT317AT
IC V RGLTR-ADJ-NEG -37/1.2V TO-220 PKG
IC OP AMP LOW-BIAS-H-IMPD 8-DIP-P PKG
IC OP AMP LOW-BIAS-H-IMPD 8-DIP-P PKG
IC OP AMP LOW-BIAS-H-IMPD 8-DIP-P PKG
IC MUXR/DATA-SEL CMOS/74HC 2-TO-1-LINE
IC V RGLTR-V-REF-FXD 1.22/1.24V TO-92
IC OP AMP LOW-OFS 8-DIP-P PKG
00000 LT337AT
3
00000 LT1012CN8
00000 LT1012CN8
00000 LT1012CN8
04713 MC74HC157N
27014 LM385BZ-1.2
06665 OP-07CP
1
1
1
VR1
W1
1902-0951
1
1
DIODE-ZNR 5.1V 5% D0-35 PD=.4W TC .035%
CABLE ASSEMBLY, SAFETY GROUND
28480 1902-0951
01141-61602
28480 01141-61602
84
Download from Www.Somanuals.com. All Manuals Search And Download.
Service
Replaceable Parts
Locator Table for Control and Power Supply
Ref. Grid Ref Grid Ref Grid Ref Grid Ref Grid Ref Grid Ref Grid Ref Grid
Des. Loc Des Loc Des Loc Des Loc Des Loc Des Loc Des Loc Des Loc
C1
C2
C3
C4
C5
C6
C7
C8
C9
C-2
F-3
H-2
J-3
C-3
C-5
B-3
C-5
C-5
C17
C-6 CR10 C-5
J1
J2
J3
G-7
B-7
C-7
R5
R6
R7
R8
R9
B-6
B-6
C-4
C-5
C-4
R22 B-3
R23 B-5
R24 B-5
R25 B-2
R26 B-2
R27 A-5
R28 A-5
R29 B-6
R30 B-6
R31 B-3
R32 B-2
R40 C-2
R41 A-2
R42 C-2
R43 C-2
R44 C-3
R45 C-3
R46 C-3
R47 C-2
R48 D-3
U5
U6
U7
U8
U9
D-6
B-3
C-3
B-3
A-5
C18 B-5 CR11 C-3
C19 B-6 CR12 C-6
C20 B-5 CR13 C-3
C21 B-5 CR14 C-6
C22 B-6 CR17 B-5
C23 B-4 CR18 B-6
C24 B-6
MP1 H-1
MP2 E-3
MP3 D-3
MP4 E-5
MP7 E-4
MP8 G-4
MP9 G-4
R10 C-5
R11 C-5
R12 C-5
R13 E-3
R14 C-4
R15 C-4
R16 C-4
R17 A-2
R18 C-6
R19 C-6
R21 B-3
RP1 B-3
U10 C-2
U11 C-3
S1
S2
H-7
E-2
C25 B-4 DS1 G-2
C26 G-6
VR1 E-3
C10 C-5
C11 E-3
C12 A-3
C13 A-3
C14 B-4
C15 A-3
C16 C-4
E1
F-3 E2
F-3
F-6
E-6
T1
H-5
CR2
CR3
P1
E-7
U1
U2
U3
U4
D-4
D-5
A-2
D-3
CR5 C-3 F1
CR6 C-5
E-6
R3
R4
A-6
A-6
CR9 C-3 HS1 D-5
85
Download from Www.Somanuals.com. All Manuals Search And Download.
Service
Replaceable Parts
86
Download from Www.Somanuals.com. All Manuals Search And Download.
Index
Numerics
100x attenuator test, 32
10x attenuator test, 30
cabling diagram, 75
A
F
ac adapter, 20
ac coupling adapters, 19
ac low frequency, 55
accessories
available to order, 8
ground leads, 14
mini grabbers, 15
probe tips, 14
G
test board, 26
connector
using, 13
adapters, 16
H
high frequency
attenuator, 61
compatibility, 18
attenuator adapter, 49
initial, 13
probe, 40, 42
altitude, 58
D
amplifiers, 23
initial adjustment, 11
assemblies, removing & replacing,
dc gain accuracy, 28
66
attenuator
adapters, 61
differential amplifiers, 23
differential input range, 55
differential probe
B
L
bandwidth
line voltage, 10
disassembly, 66
specification, 55
test, 34
blocking dc, 18
block-level theory, 59
local control, 61
reassembly, 67
locator table, 85
dimensions, 58
low frequency response, 44
direct mail order parts, 71
low-frequency path, 60
displayed noises, displayed, 55
87
Download from Www.Somanuals.com. All Manuals Search And Download.
Index
M
probe preparation, 41
making measurements, 13
manufacturer’s codes, 72
maximum input voltage, 55
R
troubleshooting
N
noises, 55
O
U
offset, 20
removing assemblies, 66
using accessories, 13
offset functions, 61
replaceable parts, 71
offset null, 11
replaceable parts list, 78
operating range, 55
ordering information, 71
output impedance, 55
overload recovery, 55
probe control & power module,
P
parts list, 71, 78
power module, 60
disassembly, 70
setup procedure, 11, 12
troubleshooting, 64
power supply, 61
probe adapter
disassembly, 68
system preparation, 10
reassembly, 68
probe adjustment, 40
probe control
disassembly, 70
troubleshooting, 64
probe gain test, 28
probe tips, 14
10x attenuator accuracy, 30
bandwidth, 34
procedure
calibration, 27
attenuator adjustment, 50
CMRR, 36
bandwidth, 34
record calibration, 39
test board, 18, 26, 62
test equipment, 9, 26
CMRR test, 36
probe adjustment, 42
88
Download from Www.Somanuals.com. All Manuals Search And Download.
commercial license terms, and
non-DOD Departments and
Agencies of the U.S. Government
will receive no greater than
Restricted Rights as defined in
FAR 52.227-19(c)(1-2) (June
1987). U.S. Government users will
receive no greater than Limited
Rights as defined in FAR 52.227-
14 (June 1987) or DFAR 252.227-
7015 (b)(2) (November 1995), as
applicable in any technical data.
Warranty
Notices
© Agilent Technologies, Inc.
2000-2004
The material contained in this
document is provided “as is,” and
is subject to being changed,
without notice, in future editions.
Further, to the maximum extent
permitted by applicable law,
Agilent disclaims all warranties,
either express or implied, with
regard to this manual and any
information contained herein,
including but not limited to the
implied warranties of
merchantability and fitness for a
particular purpose. Agilent shall
not be liable for errors or for
incidental or consequential
damages in connection with the
furnishing, use, or performance
of this document or of any
information contained herein.
Should Agilent and the user have
a separate written agreement
with warranty terms covering the
material in this document that
conflict with these terms, the
warranty terms in the separate
agreement shall control.
No part of this manual may be
reproduced in any form or by any
means (including electronic
storage and retrieval or
translation into a foreign
language) without prior
agreement and written consent
from Agilent Technologies, Inc. as
governed by United States and
international copyright laws.
Manual Safety Notices
CAUTION:
A CAUTION notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly
performed or adhered to, could
result in damage to the product or
loss of important data. Do not
proceed beyond a CAUTION
notice until the indicated
Manual Part Number
01141-97002, July 2004
Print History
01141-97000, June 2000
01141-97001, September 2002
01141-97002, July 2004
Agilent Technologies, Inc.
1900 Garden of the Gods Road
Colorado Springs, CO 80907 USA
conditions are fully understood
and met.
WARNING:
A WARNING notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly
performed or adhered to, could
result in personal injury or death.
Do not proceed beyond a
WARNING notice until the
indicated conditions are fully
understood and met.
Technology Licenses
The hardware and/or software
described in this document are
furnished under a license and
may be used or copied only in
accordance with the terms of
such license.
Restricted Rights
Legend
If software is for use in the
performance of a U.S.
Government prime contract or
subcontract, Software is delivered
and licensed as “Commercial
computer software” as defined in
DFAR 252.227-7014 (June 1995),
or as a “commercial item” as
defined in FAR 2.101(a) or as
“Restricted computer software”
as defined in FAR 52.227-19
(June 1987) or any equivalent
agency regulation or contract
clause. Use, duplication or
disclosure of Software is subject
to Agilent Technologies’ standard
Agilent Technologies
P.O. Box 2197
1900 Garden of the Gods Road
Download from Www.Somanuals.com. All Manuals Search And Download.
any unintended operation.
Safety Symbols
Product Safety
Notices
This apparatus has been designed
and tested in accordance with
IEC Publication 1010, Safety
Requirements for Measuring
Apparatus, and has been supplied
in a safe condition. This is a
Safety Class I instrument
(provided with terminal for
protective earthing). Before
applying power, verify that the
correct safety precautions are
taken (see the following
Service instructions are for
trained service personnel. To
avoid dangerous electric shock,
do not perform any service unless
qualified to do so. Do not attempt
internal service or adjustment
unless another person, capable of
rendering first aid and
!
Instruction manual symbol: the
product is marked with this
symbol when it is necessary for
you to refer to the instruction
manual in order to protect
against damage to the product.
resuscitation, is present.
Do not install substitute parts or
perform any unauthorized
modification to the instrument.
Hazardous voltage symbol.
Capacitors inside the instrument
may retain a charge even if the
instrument is disconnected from
its source of supply.
warnings). In addition, note the
external markings on the
instrument that are described
under "Safety Symbols."
Earth terminal symbol: Used to
indicate a circuit common
connected to grounded chassis.
Do not operate the instrument in
the presence of flammable gasses
or fumes. Operation of any
electrical instrument in such an
environment constitutes a definite
safety hazard.
Warnings
Before turning on the instrument,
you must connect the protective
earth terminal of the instrument
to the protective conductor of the
(mains) power cord. The mains
plug shall only be inserted in a
socket outlet provided with a
protective earth contact. You
must not negate the protective
action by using an extension cord
(power cable) without a
Do not use the instrument in a
manner not specified by the
manufacturer.
To clean the instrument
If the instrument requires
cleaning: (1) Remove power from
the instrument. (2) Clean the
external surfaces of the
protective conductor
(grounding). Grounding one
conductor of a two-conductor
outlet is not sufficient protection.
instrument with a soft cloth
dampened with a mixture of mild
detergent and water. (3) Make
sure that the instrument is
completely dry before
Only fuses with the required rated
current, voltage, and specified
type (normal blow, time delay,
etc.) should be used. Do not use
repaired fuses or short-circuited
fuseholders. To do so could cause
a shock or fire hazard.
reconnecting it to a power source.
If you energize this instrument by
an auto transformer (for voltage
reduction or mains isolation), the
common terminal must be
connected to the earth terminal of
the power source.
Whenever it is likely that the
ground protection is impaired,
you must make the instrument
inoperative and secure it against