Teledyne Carbon Monoxide Alarm 360E User Manual

INSTRUCTION MANUAL  
MODEL 360E  
CARBON DIOXIDE ANALYZER  
© TELEDYNE INSTRUMENTS  
ADVANCED POLLUTION INSTRUMENTATION DIVISION  
9480 CARROLL PARK DRIVE  
SAN DIEGO, CA 92121-5201  
USA  
Toll-free Phone: 800-324-5190  
Phone: 858-657-9800  
Fax: 858-657-9816  
Email: api-sales@teledyne.com  
Website: http://www.teledyne-api.com/  
04584 REV. C8  
Copyright 2008 Teledyne Advanced Pollution  
Instrumentation  
DCN 5341  
06 March 2009  
PRINT DATE: 06 March 2009  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual  
M360E Documentation  
SAFETY MESSAGES  
Your safety and the safety of others is very important. We have provided many important safety messages in  
this manual. Please read these messages carefully.  
A safety message alerts you to potential hazards that could hurt you or others. Each safety message is  
associated with a safety alert symbol. These symbols are found in the manual and inside the instrument. The  
definition of these symbols is described below:  
GENERAL SAFETY HAZARD: Refer to the instructions for  
details on the specific hazard.  
CAUTION: Hot Surface Warning  
CAUTION: Electrical Shock Hazard  
TECHNICIAN SYMBOL: All operations marked with this  
symbol are to be performed by qualified maintenance  
personnel only.  
CAUTION  
The analyzer should only be used for the purpose and in the manner described in this  
manual. If you use the analyzer in a manner other than that for which it was intended,  
unpredictable behavior could ensue with possible hazardous consequences.  
05232 Rev B3  
i
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual  
M360E Documentation  
TABLE OF CONTENTS  
3.3.1. Initial O2 Sensor Calibration Procedure ..................................................................................................................30  
3.3.1.1. O2 Calibration Setup .......................................................................................................................................30  
3.3.1.2. O2 Calibration Method.....................................................................................................................................30  
5.4.4. Zero/Span Valve with External CO2 Scrubber (Option 53)......................................................................................44  
5.6.1.1. Paramagnetic measurement of O2..................................................................................................................46  
05232 Rev B3  
iii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
M360E Documentation  
Model 360E Instruction Manual  
iv  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual  
M360E Documentation  
10.1.4.1. Ambient CO2 Interference Rejection ...........................................................................................................167  
05232 Rev B3  
v
Download from Www.Somanuals.com. All Manuals Search And Download.  
M360E Documentation  
Model 360E Instruction Manual  
10.3.5.2. I2C Watch Dog Circuitry ..............................................................................................................................180  
vi  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual  
M360E Documentation  
11.5.3. I2C Bus................................................................................................................................................................210  
05232 Rev B3  
vii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
M360E Documentation  
Model 360E Instruction Manual  
LIST OF APPENDICES  
APPENDIX A - VERSION SPECIFIC SOFTWARE DOCUMENTATION  
APPENDIX A-1: M360E Software Menu Trees, Revision G.4  
APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4  
APPENDIX A-3: Warnings and Test Functions, Revision G.4  
APPENDIX A-4: M360E Signal I/O Definitions, Revision G.4  
APPENDIX A-5: M360E iDAS Functions, Revision G.4  
APPENDIX A-6: Terminal Command Designators, Revision G.4  
APPENDIX B - M360E SPARE PARTS LIST  
APPENDIX C - REPAIR QUESTIONNAIRE - M360E  
APPENDIX D - ELECTRONIC SCHEMATICS  
LIST OF FIGURES  
Figure 3-13: O2 Sensor Calibration Set Up ........................................................................................................... 30  
viii  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual  
M360E Documentation  
Figure 10-4: Affect of CO2 in the Sample on CO2 MEAS & CO2 REF ............................................................... 166  
Figure 11.9: M360E – Internal Pneumatics with O2 Sensor Option 65 ............................................................... 202  
LIST OF TABLES  
05232 Rev B3  
ix  
Download from Www.Somanuals.com. All Manuals Search And Download.  
M360E Documentation  
Model 360E Instruction Manual  
Table 6-23: CO2 Concentration Alarm Default Settings ...................................................................................... 122  
User Notes  
x
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual  
M360E Documentation  
1. M360E DOCUMENTATION  
Thank you for purchasing the Model 300E Gas Filter Correlation Carbon monoxide Analyzer!  
The documentation (part number 04584) for this instrument is available in several different formats:  
Printed format, or;  
Electronic format on a CD-ROM.  
The electronic manual is in Adobe® Systems Inc. “Portable Document Format”. The Adobe® Acrobat Reader®  
software, which is necessary to view these files, can be downloaded for free from the internet at  
http://www.adobe.com/.  
The electronic version of the manual has many advantages:  
Keyword and phrase search feature  
Figures, tables and internet addresses are linked so that clicking on the item will display the associated  
feature or open the website.  
A list of chapters as well as thumbnails of each page is displayed to the left of the text.  
Entries in the table of contents are linked to the corresponding locations in the manual.  
Ability to print s (or all) of the manual  
Additional documentation for the Model 360E CO2 Analyzer is available from Teledyne Instruments’ website at  
APICOM software manual, part number 03945  
Multi-drop manual, part number 01842  
DAS Manual, part number 02837.  
1.1. Using This Manual  
This manual has the following data structures:  
1.0 TABLE OF CONTENTS:  
Outlines the contents of the manual in the order the information is presented. This is a good overview of the  
topics covered in the manual. There is also a list of tables, a list of figures and a list of appendices. In the  
electronic version of the manual, clicking on any of these table entries automatically views that section.  
2.0 SPECIFICATIONS AND WARRANTY  
This section contains a list of the analyzer’s performance specifications, a description of the conditions and  
configuration under which EPA equivalency was approved and Teledyne Instruments Incorporated’s warranty  
statement.  
3.0 GETTING STARTED:  
A concise set of instructions for setting up, installing and running your analyzer for the first time.  
4.0 FAQ:  
Answers to the most frequently asked questions about operating the analyzer.  
05232 Rev B3  
1
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
M360E Documentation  
Model 360E Instruction Manual  
5.0 OPTIONAL HARDWARE & SOFTWARE  
A description of optional equipment to add functionality to your analyzer.  
6.0 OPERATION INSTRUCTIONS  
This section includes step-by-step instructions for operating the analyzer and using its various features and  
functions.  
7.0 CALIBRATION PROCEDURES  
General information and step by step instructions for calibrating your analyzer.  
8.0 EPA PROTOCOL CALIBRATION  
Because CO2 is not declared a criteria air pollutant by the US EPA, EPA equivalency is not required for this type  
of analyzer. Therefore no special calibration methods are needed to satisfy EPA requirements.  
9.0 INSTRUMENT MAINTENANCE  
Description of certain preventative maintenance procedures that should be regularly performed on you  
instrument to keep it in good operating condition. This section also includes information on using the iDAS to  
record diagnostic functions useful in predicting possible component failures before they happen.  
10.0 THEORY OF OPERATION  
An in-depth look at the various principals by which your analyzer operates as well as a description of how the  
various electronic, mechanical and pneumatic components of the instrument work and interact with each other.  
A close reading of this section is invaluable for understanding the instrument’s operation.  
11.0 TROUBLESHOOTING:  
This section includes pointers and instructions for diagnosing problems with the instrument, such as excessive  
noise or drift, as well as instructions on performing repairs of the instrument’s major subsystems.  
APPENDICES:  
For easier access and better updating, some information has been separated out of the manual and placed in a  
series of appendices at the end of this manual. These include: software menu trees, warning messages,  
definitions of iDAS & serial I/O variables, spare parts list, repair questionnaire, interconnect listing and drawings,  
and electronic schematics.  
NOTE  
Throughout this manual, words printed in capital, bold letters, such as SETUP or ENTR  
represent messages as they appear on the analyzer’s front panel display.  
NOTE  
The flowcharts in this manual contain typical representations of the analyzer’s display  
during the various operations being described. These representations are not intended  
to be exact and may differ slightly from the actual display of your instrument.  
User Notes  
2
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual  
Specifications, Approvals and Warranty  
2. SPECIFICATIONS, APPROVALS AND WARRANTY  
2.1. Specifications  
Table 2-1: Model 360E Basic Unit Specifications  
Min/Max Range  
In 1ppb increments from 50ppb to 2 000ppm, dual ranges or auto ranging  
(Physical Analog Output)  
Measurement Units  
Zero Noise  
ppb, ppm, µg/m3, mg/m3, %(user selectable)  
< 0.1 ppm (RMS)  
Span Noise  
< 1% of reading (RMS)  
< 0.2 ppm1  
<0.25 ppm1  
<0.5 ppm1  
1% of reading above 50 PPM1  
Lower Detectable Limit1  
Zero Drift (24 hours)  
Zero Drift (7 days)  
Span Drift (7 Days)  
Linearity  
1% of full scale  
Precision  
0.5% of reading  
Temperature Coefficient  
Voltage Coefficient  
Lag Time  
< 0.1% of Full Scale per oC  
< 0.05% of Full Scale per V  
10 sec  
Rise/Fall Time  
95% in <60 sec  
Sample Flow Rate  
800cm3/min. ±10%  
O2 Sensor option adds 120 cm³/min to total flow though when installed;  
5-40oC  
Temperature Range  
Humidity Range  
0 - 95% RH, non-condensing  
7" x 17" x 23.5" (178 mm x 432 mm x 597 mm)  
38 lbs. (17 kg); add 1 lbs (0.5 kg) for IZS  
Dimensions H x W x D  
Weight, Analyzer  
AC Power Rating  
100 V, 50/60 Hz (3.25A);  
115 V, 60 Hz (3.0 A);  
220 – 240 V, 50/60 Hz (2.5 A)  
Environmental  
Installation category (over-voltage category) II; Pollution degree 2  
Three (3) Outputs  
Analog Outputs  
Analog Output Ranges  
0.1V, 1 V, 5 V, 10 V, 2-20 or 4-20 mA isolated current loop.  
All Ranges with 5% Under/Over Range  
Analog Output Resolution  
Status Outputs  
Control Inputs  
1 part in 4096 of selected full-scale voltage  
8 Status outputs - opto-isolated; including 2 alarm outputs  
6 Control Inputs, 3 defined, 3 spare  
Serial I/O  
One (1) RS-232; One (1) RS-485  
Baud Rate : 300 – 115200: Optional Ethernet Interface  
Alarm outputs  
Certifications  
2 opto-isolated alarms outputs with user settable alarm limits  
CE: EN61010-1:90 + A1:92 + A2:95, EN61326 - Class A  
1 At constant temperature and voltage.  
05232 Rev B3  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Specifications, Approvals and Warranty  
Model 360E Instruction Manual  
2.2. CE Mark Compliance  
Emissions Compliance  
The Teledyne Instruments Model 360E Gas Filter Correlation CO2 Analyzer was tested and found to be fully  
compliant with:  
EN61326 (1997 w/A1: 98) Class A, FCC Part 15 Subpart B section 15.107 Class A, ICES-003 Class A (ANSI  
C63.4 1992) & AS/NZS 3548 (w/A1 & A2; 97) Class A.  
Tested on 11-29-2001 at CKC Laboratories, Inc., Report Number CE01-249.  
Safety Compliance  
The Teledyne Instruments Model 360E Gas Filter Correlation CO2 Analyzer was tested and found to be fully  
compliant with:  
IEC 61010-1:90 + A1:92 + A2:95,  
Tested on 02-06-2002 at Nemko, Report Number 2002-012219.  
2.3. Warranty  
WARRANTY POLICY (02024D)  
Prior to shipment, T-API equipment is thoroughly inspected and tested. Should equipment failure occur, T-API  
assures its customers that prompt service and support will be available.  
COVERAGE  
After the warranty period and throughout the equipment lifetime, T-API stands ready to provide on-site or in-plant  
service at reasonable rates similar to those of other manufacturers in the industry. All maintenance and the first  
level of field troubleshooting is to be performed by the customer.  
NON-API MANUFACTURED EQUIPMENT  
Equipment provided but not manufactured by T-API is warranted and will be repaired to the extent and according  
to the current terms and conditions of the respective equipment manufacturer’s warranty.  
GENERAL  
During the warranty period, T-API warrants each Product manufactured by T-API to be free from defects in  
material and workmanship under normal use and service. Expendable parts are excluded.  
If a Product fails to conform to its specifications within the warranty period, API shall correct such defect by, in  
API's discretion, repairing or replacing such defective Product or refunding the purchase price of such Product.  
The warranties set forth in this section shall be of no force or effect with respect to any Product: (i) that has been  
altered or subjected to misuse, negligence or accident, or (ii) that has been used in any manner other than in  
accordance with the instruction provided by T-API, or (iii) not properly maintained.  
4
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Model 360E Instruction Manual  
Specifications, Approvals and Warranty  
THE WARRANTIES SET FORTH IN THIS SECTION AND THE REMEDIES THEREFORE ARE EXCLUSIVE  
AND IN LIEU OF ANY IMPLIED WARRANTIES OF MERCHANTABILITY, FITNESS FOR PARTICULAR  
PURPOSE OR OTHER WARRANTY OF QUALITY, WHETHER EXPRESSED OR IMPLIED. THE REMEDIES  
SET FORTH IN THIS SECTION ARE THE EXCLUSIVE REMEDIES FOR BREACH OF ANY WARRANTY  
CONTAINED HEREIN. API SHALL NOT BE LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL  
DAMAGES ARISING OUT OF OR RELATED TO THIS AGREEMENT OF T-API'S PERFORMANCE  
HEREUNDER, WHETHER FOR BREACH OF WARRANTY OR OTHERWISE  
Terms and Conditions  
All units or components returned to Teledyne Instruments Incorporated should be properly packed for handling  
and returned freight prepaid to the nearest designated Service Center. After the repair, the equipment will be  
returned, freight prepaid.  
User Notes  
05232 Rev B3  
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual  
Getting Started  
3. GETTING STARTED  
3.1. Unpacking and Initial Set Up  
CAUTION  
To avoid personal injury, always use two persons to lift and carry the Model 360E.  
1. Verify that there is no apparent external shipping damage. If damage has occurred, please advise the  
shipper first, then Teledyne Instruments.  
2. Included with your analyzer is a printed record of the final performance characterization performed on  
your instrument at the factory. This record, titled Final Test and Validation Data Sheet (P/N 04596) is an  
important quality assurance and calibration record for this instrument. It should be placed in the quality  
records file for this instrument.  
3. Carefully remove the top cover of the analyzer and check for internal shipping damage.  
Remove the set-screw located in the top, center of the Front panel.  
Remove the 2 screws fastening the top cover to the unit (one per side towards the rear).  
Slide the cover backwards until it clears the analyzer’s front bezel.  
Lift the cover straight up.  
NOTE  
Some versions of the 360E CO2 Analyzer may have a spring loaded fastener at the top  
center of the rear panel and as many as eight screws (four per side) fastening the top  
cover to the chassis.  
NOTE  
Printed circuit assemblies (PCAs) are sensitive to electro-static discharges too small to  
be felt by the human nervous system. Failure to use ESD protection when working with  
electronic assemblies will void the instrument warranty.  
See Chapter 12 for more information on preventing ESD damage.  
CAUTION  
Never disconnect PCAs, wiring harnesses or electronic subassemblies while under  
power.  
4. Inspect the interior of the instrument to make sure all circuit boards and other components are in good  
shape and properly seated.  
05232 Rev B3  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Getting Started  
Model 360E Instruction Manual  
5. Check the connectors of the various internal wiring harnesses and pneumatic hoses to make sure they  
are firmly and properly seated.  
6. Verify that all of the optional hardware ordered with the unit has been installed. These are listed on the  
paperwork accompanying the analyzer.  
7. VENTILATION CLEARANCE: Whether the analyzer is set up on a bench or installed into an instrument  
rack, be sure to leave sufficient ventilation clearance.  
AREA  
MINIMUM REQUIRED CLEARANCE  
Back of the instrument  
4 in.  
1 in.  
1 in.  
Sides of the instrument  
Above and below the instrument  
Various rack mount kits are available for this analyzer. See Section _Rack_Mount_Option5.1 of this  
manual for more information.  
8
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 360E Instruction Manual  
Getting Started  
3.1.1. M360 Layout  
Figure 3-1 shows the front panel layout of the analyzer. Figure 3-2 shows the rear panel configuration with  
optional zero-air scrubber mounted to it and two optional fittings for the IZS option.  
Figure 3-3 shows a top-down view of the analyzer. The shown configuration includes the Ethernet board, IZS  
option, zero-air scrubber and an additional sample dryer. See Chapter 5 for optional equipment.  
MODE FIELD  
MESSAGE FIELD  
CONCENTRATION FIELD  
STATUS LED’s  
LOCKING SCREW  
FASTENER  
FASTENER  
SAMPLE  
CAL  
CO2 =  
RANGE = 500.0  
<TST TST> CAL  
SAMPLE A  
SETUP  
FAULT  
POWER  
GAS FILTER CORRELATION CO2 ANALYZER- MODEL 360E  
KEY DEFINITIONS  
KEYBOARD  
ON / OFF SWITCH  
Figure 3-1: Front Panel Layout  
Figure 3-2: Rear Panel Layout  
05232 Rev B3  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Getting Started  
Model 360E Instruction Manual  
Front Panel  
IR Source  
On/Off Switch &  
Circuit Breaker  
Particulate Filter  
GFC Wheel Housing  
Purge Gas Inlet &  
Flow Control Orifice  
Optical Bench Gas  
Inlet  
Optional  
Sample/Cal  
Valve  
GFC Motor  
GFC Wheel Housing &  
IR Source Heat Sync  
Sample Gas  
Critical Flow  
Orifice  
Optional  
Zero/Span  
Valve  
Gas Flow  
Sensor Assy  
Flow Sensor  
Optional  
Shutoff  
Valve  
Purge Gas  
Pressure Control  
Assy  
Pump Assy  
Sample Gas  
Pressure Sensor  
Optical Bench Gas  
Outlet  
Sample Gas  
Temperature  
Sensor  
Optional  
Ethernet Card  
CPU Card  
Mother  
Board  
Power  
Receptacle  
Rear Panel  
Fan  
Figure 3-3: Assembly Layout  
10  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 360E Instruction Manual  
Getting Started  
Sample Gas Outlet  
fitting  
Sample Gas Flow  
Sensor  
Sample Chamber  
Sync/Demod PCA  
Housing  
Pressure Sensor(s)  
Bench  
Temperature  
Thermistor  
Shock Absorbing  
Mounting Bracket  
Opto-Pickup  
PCA  
Purge Gas  
Pressure Regulator  
IR Source  
GFC Wheel  
Heat Sync  
GFC Wheel Motor  
GFC Temperature  
Sensor  
Purge Gas  
Inlet  
GFC Heater  
Figure 3-4: Optical Bench Layout  
Figure 3-5: M360E Internal Gas Flow  
05232 Rev B3  
11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Getting Started  
Model 360E Instruction Manual  
3.1.2. Electrical Connections  
3.1.2.1. Power Connection  
Attach the power cord to the analyzer and plug it into a power outlet capable of carrying at least 10 A current at  
your AC voltage and that it is equipped with a functioning earth ground.  
CAUTION  
Check the voltage and frequency label on the rear panel of the instrument  
(See Figure 3-2 for compatibility with the local power before plugging the  
M360E into line power.  
Do not plug in the power cord if the voltage or  
frequency is incorrect.  
CAUTION  
Power connection must have functioning ground connection.  
Do not defeat the ground wire on power plug.  
Turn off analyzer power before disconnecting or  
connecting electrical subassemblies.  
Do not operate with cover off.  
The M360E analyzer can be configured for both 100-130 V and 210-240 V at either 50 or 60 Hz. To avoid  
damage to your analyzer, make sure that the AC power voltage matches the voltage indicated on the rear panel  
serial number label and that the frequency is between 47 and 63 Hz.  
12  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Model 360E Instruction Manual  
Getting Started  
3.1.2.2. Output Connections  
The M360E is equipped with several analog output channels accessible through a connector on the back panel  
of the instrument. The standard configuration for these outputs is mVDC. An optional current loop output is  
available for each.  
When the instrument is in its default configuration, channels A1 and A2 output a signal that is proportional to the  
CO2 concentration of the sample gas. Either can be used for connecting the analog output signal to a chart  
recorder or for interfacing with a data logger.  
Output A3 is only used on the Model 306E if the optional O2 sensor is installed.  
Channel A4 is special. It can be set by the user (see Section 6.13.9) to output any one of the parameters  
accessible through the <TST TST> keys of the units sample display.  
To access these signals attach a strip chart recorder and/or data-logger to the appropriate analog output  
connections on the rear panel of the analyzer.  
ANALOG  
A1  
A2  
A3  
A4  
1
2
3
4
5
6
7
8
Pin-outs for the analog output connector at the rear panel of the instrument are:  
Table 3-1: M360E Analog Output Pin Outs  
PIN  
1
ANALOG OUTPUT  
A1  
VDC SIGNAL  
V Out  
MADC SIGNAL  
I Out +  
2
Ground  
V Out  
I Out -  
3
I Out +  
A2  
4
Ground  
V Out  
I Out -  
7
I Out +  
A3  
(Only used if O2 sensor  
is installed)  
8
Ground  
I Out -  
5
6
V Out  
I Out +  
I Out -  
A4  
Ground  
The default analog output voltage setting of the 360E CO2 Analyzer is 0 – 5 VDC with a range of 0 –  
500 ppm.  
TO change these settings, see Sections 6.13.4 and 6.8 respectively.  
05232 Rev B3  
13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Getting Started  
Model 360E Instruction Manual  
3.1.2.3. Connecting the Status Outputs  
If you wish utilize the analyzer’s status outputs to interface with a device that accepts logic-level digital inputs,  
such as programmable logic controllers (PLC’s) they are accessed via a 12-pin connector on the analyzer’s rear  
panel labeled STATUS.  
STATUS  
1
2
3
4
5
6
7
8
D
+
NOTE  
Most PLC’s have internal provisions for limiting the current the input will draw. When  
connecting to a unit that does not have this feature, external resistors must be used to  
limit the current through the individual transistor outputs to 50mA (120 Ω for 5V  
supply).  
The pin assignments for the status outputs can be found in the table below:  
Table 3-2: M360E Status Output Pin Outs  
STATUS  
DEFINITION  
OUTPUT #  
CONDITION  
1
2
SYSTEM OK  
On if no faults are present.  
On if CO2 concentration measurement is valid.  
CONC VALID  
If the CO2 concentration measurement is invalid, this bit is OFF.  
On if unit is in high range of DUAL or AUTO range modes.  
On whenever the instruments ZERO point is being calibrated.  
On whenever the instruments SPAN point is being calibrated.  
On whenever the instrument is in DIAGNOSTIC mode.  
3
4
5
6
HIGH RANGE  
ZERO CAL  
SPAN CAL  
DIAG MODE  
On whenever the measured CO2 concentration is above the set point for  
ALM1  
7
8
ALARM1  
ALARM2  
On whenever the measured CO2 concentration is above the set point for  
ALM2  
D
+
EMITTER BUSS  
DC POWER  
The emitters of the transistors on pins 1-8 are bussed together.  
+ 5 VDC  
Digital Ground  
The ground level from the analyzer’s internal DC power supplies.  
14  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Model 360E Instruction Manual  
Getting Started  
3.1.2.4. Connecting the Control Inputs  
If you wish to use the analyzer to remotely activate the zero and span calibration modes, several digital control  
inputs are provided through a 10-pin connector labeled CONTROL IN on the analyzer’s rear panel.  
There are two methods for energizing the control inputs. The internal +5V available from the pin labeled “+” is  
the most convenient method. However, if full isolation is required, an external 5 VDC power supply should be  
used.  
CONTROL IN  
CONTROL IN  
A
B
C
D
E
F
U
+
A
B
C
D
E
F
U
+
Z
E
R
O
S
P
A
N
Z
E
R
O
S
P
A
N
5 VDC Power  
Supply  
+
-
External Power Connections  
Local Power Connections  
The pin assignments for the digital control inputs can be found in the table below:  
Table 3-3: M360E Control Input Pin Outs  
STATUS  
DEFINITION  
INPUT #  
ON CONDITION  
REMOTE ZERO  
CAL  
The Analyzer is placed in Zero Calibration mode. The mode field of  
the display will read ZERO CAL R.  
A
B
REMOTE  
SPAN CAL  
The Analyzer is placed in Span Calibration mode. The mode field of  
the display will read SPAN CAL R.  
SPARE  
SPARE  
C
D
E
F
SPARE  
SPARE  
Digital Ground  
May be connected to the ground of the data logger/recorder.  
Input pin for +5 VDC required to activate pins A – F. This can be from  
an external source or from the “+” pin of the instruments STATUS  
connector.  
Pull-up supply for  
inputs  
U
Internal +5V  
Supply  
Internal source of +5V which can be used to actuate control inputs  
when connected to the U pin.  
+
05232 Rev B3  
15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Getting Started  
Model 360E Instruction Manual  
3.1.2.5. Connecting the Serial Ports  
If you wish to utilize either of the analyzer’s two serial interface COMM ports, refer to Section 6.11 of this manual  
for instructions on their configuration and usage.  
3.1.2.6. Connecting to a LAN or the Internet  
If your unit has a Teledyne Instruments Ethernet card (Option 63), plug one end into the 7’ CAT5 cable supplied  
with the option into the appropriate place on the back of the analyzer (see Figure 5-5 in Section 5.5.3) and the  
other end into any nearby Ethernet access port.  
NOTE:  
The M360E firmware supports dynamic IP addressing or DHCP.  
If your network also supports DHCP, the analyzer will automatically configure its LAN  
connection appropriately,  
If your network does not support DHCP, see Section 6.11.6.3 for instructions on  
manually configuring the LAN connection.  
3.1.2.7. Connecting to a Multidrop Network  
If your unit has a Teledyne Instruments RS-232 multidrop card (Option 62), see Section 6.11.7 for instructions  
on setting it up.  
3.1.3. Pneumatic Connections:  
3.1.3.1. Basic Pneumatic Connections  
CAUTION  
In order to prevent dust from getting into the gas flow channels of your analyzer, it was  
shipped with small plugs inserted into each of the pneumatic fittings on the back panel.  
Make sure that all of these dust plugs are removed before attaching  
exhaust and supply gas lines.  
Figures 3-3 and 3-4 illustrate the most common configurations for gas supply and exhaust lines to the Model  
360E Analyzer. Figure 3-5 illustrates the internal gas flow of the instrument in its basic configuration.  
Please refer to Figure 3-2 for pneumatic connections at the rear panel and Table 3-4 for nomenclature.  
NOTE  
Sample and calibration gases should only come into contact with PTFE (Teflon), FEP,  
glass, stainless steel or brass.  
16  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Model 360E Instruction Manual  
Table 3-4: Model 360E Rear Panel Pneumatic Connections  
Getting Started  
REAR PANEL LABEL  
SAMPLE  
FUNCTION  
Connect a gas line from the source of sample gas here.  
Calibration gasses are also inlet here on units without zero/span/shutoff valve  
or IZS options installed.  
EXHAUST  
Connect an exhaust gas line of not more than 10 meters long here.  
On units with zero/span/shutoff valve options installed, connect a gas line to the  
source of calibrated span gas here.  
PRESSURE SPAN  
Span gas vent outlet for units with zero/span/shutoff valve options installed.  
Connect an exhaust gas line of not more than 10 meters long here.  
VENT SPAN  
IZS  
Internal zero air scrubber.  
On units with zero/span/shutoff valve options installed but NO internal zero  
air scrubber, attach a gas line to the source of zero air here.  
This inlet supplies purge air to the GFC wheel housing (see Section 10.2.3)  
Connect a source of dried air that has been scrubbed of CO2.  
PURGE IN  
Figure 3-6: Pneumatic Connections–Basic Configuration–Using Bottled Span Gas  
05232 Rev B3  
17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Getting Started  
Model 360E Instruction Manual  
Figure 3-7: Pneumatic Connections–Basic Configuration–Using Gas Dilution Calibrator  
1. Attach a sample inlet line to the sample inlet port. The SAMPLE input line should not be more than 2  
meters long.  
NOTE  
Ideally, the pressure of the sample gas should be at ambient pressure (0 psig).  
Maximum pressure of sample gas should not exceed 1.5 in-Hg over ambient.  
In applications where the sample gas is received from a pressurized manifold, a vent  
must be placed as shown to equalize the sample gas with ambient atmospheric  
pressure before it enters the analyzer.  
This vent line must be:  
At least 0.2m long  
No more than 2m long and vented outside the shelter or immediate area  
surrounding the instrument.  
2. Attach sources of zero air and span gas (see Figures 3-3 through 3-8 inclusive).  
Span Gas is a gas specifically mixed to match the chemical composition of the type of gas being  
measured at near full scale of the desired measurement range.  
In the case of CO2 measurements made with the Teledyne Instruments Model 360E Analyzer it is  
recommended that you use a gas calibrated to have a CO2 content equaling 80% of the range of  
compositions being measured.  
18  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 360E Instruction Manual  
Getting Started  
EXAMPLE: If the application is to measure between 0 ppm and 500 ppm, an appropriate Span Gas  
would be 400 ppm. If the application is to measure between 0 ppm and 100 ppm, an appropriate Span  
Gas would be 80 ppm.  
Span Gas can be purchased in pressurized canisters or created using Dynamic Dilution Calibrator  
such as the Teledyne Instruments Model 700 and a source of dried air scrubbed of CO2 such as a  
Teledyne Instruments Model 701 Zero Air Generator in combination with a canister of indicating  
soda lime (such as Teledyne Instruments P/N 037600000).  
Zero Air is similar in chemical composition to the Earth’s atmosphere but scrubbed of all  
components that might affect the analyzer’s readings.  
In the case of CO2 measurements this means less than 0.1 ppm of CO2 and Water Vapor (when  
dew point <-15˚C). Zero Air can be purchased in pressurized canisters or created using a Teledyne  
Instruments Model 701 Zero Air Generator in combination with a canister of indicating soda lime  
(such as Teledyne Instruments P/N 037600000).  
3. Attach an exhaust line to the exhaust outlet port.  
The exhaust from the pump and vent lines should be vented to atmospheric pressure using  
maximum of 10 meters of ¼” PTEF tubing.  
CAUTION  
Venting should be outside the shelter or immediate area surrounding the instrument.  
4. Attach a source of dried air scrubbed of CO2 to the purge inlet port.  
NOTE  
The minimum gas pressure of the source of purge air should be 7.5 psig.  
If the source of the purge air is shared by a Teledyne Instruments M700 (as shown in  
figure 3-7) the minimum gas pressure should be 25 psig and should not exceed 35 psig.  
5. Once the appropriate pneumatic connections have been made, check all pneumatic fittings for leaks  
using a procedure similar to that defined in Section 9.3.3.  
05232 Rev B3  
19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Getting Started  
Model 360E Instruction Manual  
3.1.3.2. Connections with Internal Valve Options Installed  
Figures 3-8 through 3-11 show the proper pneumatic connections for M360E’s with various optional internal  
valve sets installed.  
Figure 3-8: Pneumatic Connections–M360E with Zero/Span/Shutoff Valves (OPT 50)  
VENT here if input  
is pressurized  
Source of  
SAMPLE GAS  
Removed during  
calibration  
SAMPLE  
EXHAUST  
VENT  
VENT SPAN  
PRESSURE SPAN  
IZS  
MODEL  
360E  
Calibrated  
CO2 Gas  
at span gas  
concentration  
External  
Zero Air  
Scrubber  
PURGE LINE  
MODEL 701  
Zero Gas  
Indicating  
Soda Lime  
Generator  
Figure 3-9: Pneumatic Connections–M360E with Zero/Span/Shutoff Valves and External Zero Air  
Scrubber (OPT 51)  
20  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Model 360E Instruction Manual  
Getting Started  
Figure 3-10: Pneumatic Connections–M360E with Zero/Span Valves (OPT 52)  
VENT here if input  
is pressurized  
Source of  
SAMPLE GAS  
Removed during  
calibration  
Calibrated  
CO2 Gas  
at span gas  
Model 700 gas  
Dilution  
concentration  
Calibrator  
SAMPLE  
EXHAUST  
VENT SPAN  
PRESSURE SPAN  
IZS  
MODEL  
X00E  
External  
Zero Air  
Scrubber  
PURGE LINE  
MODEL 701  
Zero Gas  
Generator  
Indicating  
Soda Lime  
Figure 3-11: Pneumatic Connections–M360E with Zero/Span Valves with External Zero air Scrubber  
(OPT 53)  
05232 Rev B3  
21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Getting Started  
Model 360E Instruction Manual  
3.1.3.3. Pneumatic Connections to M360E in Multipoint Calibration Applications  
Some applications may require multipoint calibration checks where span gas of several different concentrations  
is needed. We recommend using high-concentration, certified, calibration gas supplied to the analyzer through a  
Gas Dilution Calibrator such as a Teledyne Instruments Model 700. This type of calibrator precisely mixes Span  
Gas and Zero Air to produce any concentration level between 0 ppm and the concentration of the calibrated gas.  
Figure 3-12 depicts the pneumatic set up in this sort of application of a Model 360E CO2 Analyzer with  
zero/span/shutoff valve option 50 installed (a common configuration for this type of application).  
VENT here if input  
is pressurized  
Source of  
SAMPLE GAS  
Removed during  
calibration  
Calibrated  
CO2 Gas  
at span gas  
Model 700 gas  
Dilution  
concentration  
Calibrator  
SAMPLE  
EXHAUST  
Gas Pressure  
VENT SPAN  
PRESSURE SPAN  
IZS  
MODEL  
X00E  
Should be  
Regulated at  
30 – 35 psig  
External  
Zero Air  
Scrubber  
PURGE LINE  
MODEL 701  
Zero Gas  
Generator  
Indicating  
Soda Lime  
Figure 3-12: Example of Pneumatic Set up for Multipoint Calibration of M360  
3.1.4. Setting the internal Purge Air Pressure.  
In order to maintain proper purge air flow though the GFC wheel hosing a manually adjustable pressure  
regulator is provided (see Figures 3-3 and 3-5). This regulator includes two output ports. One is used to supply  
purge air to the GFC wheel. The other may be used to attach a pressure gauge.  
To adjust the internal purge air pressure of the M360E:  
1. Turn off the instrument.  
2. Remove the source of zero air attached to the purge line inlet port at the back of the analyzer.  
3. Remove the analyzer’s the top cover.  
4. Remove the cap from the second, unused, output port on the pressure regulator.  
5. Attach a pressure gauge capable of measuring in the 5-10 psig range with 0.5 psig resolution to the port.  
6. Turn the instrument on.  
7. Make sure the zero air supply to the analyzer’s purge line inlet is supplying gas at a stable pressure  
above 7.5 psig.  
8. Adjust the M360E’s pressure regulator until the attached gauge reads 7.5 psig.  
9. Turn off the instrument.  
10. Remove the source of zero air attached to the purge line inlet port at the back of the analyzer.  
11. Remove the pressure gauge and reattach the end cap removed in step 4 above.  
12. Replace the analyzer’s top cover.  
22  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Model 360E Instruction Manual  
Getting Started  
3.2. Initial Operation  
If you are unfamiliar with the M360E theory of operation, we recommend that you read  
Chapter 10. For information on navigating the analyzer’s software menus, see the menu trees described in  
Appendix A.1.  
NOTE  
The analyzer’s cover must be installed to ensure that the temperatures of the GFC  
wheel and absorption cell assemblies are properly controlled.  
3.2.1. Startup  
After electrical and pneumatic connections are made, turn on the instrument. The pump, exhaust fan and PMT  
cooler fan should start immediately. The display should immediately display a single, horizontal dash in the  
upper left corner of the display. This will last approximately 30 seconds while the CPU loads the operating  
system.  
Once the CPU has completed this activity it will begin loading the analyzer firmware and configuration data.  
During this process, string of messages will appear on the analyzer’s front panel display:  
System waits 3 seconds  
then automatically begins its  
initialization routine.  
No action required.  
SELECT START OR REMOTE  
START  
:
3
.
CHECKING FLASH STATUS  
:
1
System is checking the format of  
the instrument’s flash memory  
chip.  
If at this point,  
STARTING INSTRUMENT CODE  
STARTING INSTRUMENT W/FLASH  
:
1
**FLASH FORMAT INVALID**  
appears, contact T–API customer service  
The instrument is loading  
configuration and calibration  
data from the flash chip  
:
1
The instrument is loading  
the analyzer firmware.  
M360E CO2 ANALYZER  
BOOT PROGRESS [XXXXX 50%_ _ _ _ _]  
The revision level of the  
firmware installed in your  
analyzer is briefly displayed  
SOFTWARE REVISION D.6  
BOOT PROGRESS [XXXXXXXX 80% _ _]  
SAMPLE  
TEST  
SYSTEM RESET  
CO2=X.XXX  
Firmware fully  
booted  
CAL  
CLR SETUP  
Press CLR to clear initial  
warning messages.  
(see Section 3.2.3)  
The analyzer should automatically switch to SAMPLE mode after completing the boot-up sequence and start  
monitoring CO2 gas.  
05232 Rev B3  
23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Getting Started  
Model 360E Instruction Manual  
3.2.2. Warm Up  
The M360E requires about 30 minutes warm-up time before reliable CO2 measurements can be taken. During  
that time, various portions of the instrument’s front panel will behave as follows. See Figure 3-1 for locations.  
Table 3-5: Front Panel Display During System Warm-Up  
Name  
Color  
Behavior  
Significance  
Concentration  
Field  
N/A  
Displays current,  
compensated CO2  
Concentration  
N/A  
Mode Field  
N/A  
Displays blinking  
“SAMPLE”  
Instrument is in sample mode but is still in the process  
of warming up.  
STATUS LED’s  
Sample  
Green  
On  
Unit is operating in sample mode; front panel display  
is being updated.  
Flashes On/Off when adaptive filter is active  
The instrument’s calibration is not enabled.  
Cal  
Yellow  
Red  
Off  
Fault  
Blinking  
The analyzer is warming up and hence out of specification  
for a fault-free reading. Various warning messages will  
appear.  
3.2.3. Warning Messages  
Because internal temperatures and other conditions may be outside be specified limits during the analyzer’s  
warm-up period, the software will suppress most warning conditions for 30 minutes after power up. If warning  
messages persist after the 30 minutes warm up period is over, investigate their cause using the troubleshooting  
guidelines in Chapter 11 of this manual.  
To view and clear warning messages, press:  
SAMPLE  
HVPS WARNING  
CAL MSG  
CO2 = 0.00  
TEST deactivates warning  
TEST  
CLR SETUP  
messages  
MSG activates warning  
SAMPLE  
RANGE=500.000 PPM  
MSG  
CO2 = 0.00  
messages.  
<TST TST> keys replaced with  
< TST TST > CAL  
CLR SETUP  
TEST key  
SAMPLE  
HVPS WARNING  
CO2 = 0.00  
Press CLR to clear the current  
message.  
TEST  
CAL  
MSG  
CLR SETUP  
NOTE:  
If more than one warning is active, the  
next message will take its place  
If the warning message persists  
after several attempts to clear it,  
the message may indicate a  
real problem and not an artifact  
of the warm-up period  
Once the last warning has been  
cleared, the analyzer returns to  
SAMPLE mode  
Make sure warning messages are  
not due to real problems.  
Table 3-6 lists brief descriptions of the warning messages that may occur during start up.  
24  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Model 360E Instruction Manual  
Table 3-6: Possible Warning Messages at Start-Up  
Getting Started  
MESSAGE  
MEANING  
The instrument’s A/D circuitry or one of its analog outputs is not calibrated.  
The Temperature of the optical bench is outside the specified limits.  
Remote span calibration failed while the dynamic span feature was set to turned on  
Remote zero calibration failed while the dynamic zero feature was set to turned on  
Configuration was reset to factory defaults or was erased.  
Concentration alarm 1 is enabled and the measured CO2 level is the set point.  
Concentration alarm 2 is enabled and the measured CO2 level is the set point.  
Configuration storage was reset to factory configuration or erased.  
iDAS data storage was erased.  
ANALOG CAL WARNING  
BENCH TEMP WARNING  
BOX TEMP WARNING  
CANNOT DYN SPAN  
CANNOT DYN ZERO  
CONC ALRM1 WARNING  
CONC ALRM2 WARNING  
CONFIG INITIALIZED  
DATA INITIALIZED  
O2 sensor cell temperature outside of warning limits.  
O2 CELL TEMP WARN  
PHOTO TEMP WARNING  
REAR BOARD NOT DET  
RELAY BOARD WARN  
SAMPLE FLOW WARN  
SAMPLE PRESS WARN  
SAMPLE TEMP WARN  
SOURCE WARNING  
The temperature of the IR photometer is outside the specified limits.  
The CPU is unable to communicate with the motherboard.  
The firmware is unable to communicate with the relay board.  
The flow rate of the sample gas is outside the specified limits.  
Sample gas pressure outside of operational parameters.  
The temperature of the sample gas is outside the specified limits.  
The IR source may be faulty.  
The computer was rebooted.  
SYSTEM RESET  
The Gas Filter Correlation wheel temperature is outside the specified limits.  
WHEEL TEMP WARNING  
05232 Rev B3  
25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Getting Started  
Model 360E Instruction Manual  
3.2.4. Functional Check  
1. After the analyzer’s components has warmed up for at least 30 minutes, verify that the software properly  
supports any hardware options that were installed.  
2. Check to make sure that the analyzer is functioning within allowable operating parameters. Appendix C  
includes a list of test functions viewable from the analyzer’s front panel as well as their expected values.  
These functions are also useful tools for diagnosing performance problems with your analyzer  
(Section11.1.2). The enclosed Final Test and Validation Data sheet (part number 04307) lists these  
values before the instrument left the factory.  
To view the current values of these parameters press the following key sequence on the analyzer’s front  
panel. Remember until the unit has completed its warm up these parameters may not have stabilized.  
SAMPLE  
RANGE = 500.000 PPM  
CO2 = XXX.X  
SETUP  
< TST TST > CAL  
RANGE  
RANGE11  
RANGE21  
02 RANGE2  
STABIL  
Toggle <TST TST> keys to  
scroll through list of functions  
MEAS  
REF  
MR RATIO  
PRES  
Refer to  
Table 6-2 for  
definitions of  
these test  
SAMP FL  
SAMP TEMP  
BENCH TEMP  
WHEEL TEMP  
BOX TEMP  
PHT DRIVE  
SLOPE  
functions.  
1 Only appears instrument is set  
for DUAL or AUTO reporting  
range modes.  
OFFSET  
TEST  
TIME  
2 Only appears if 02 Sensor  
Option is installed.  
3. If your analyzer has an Ethernet card (Option 63) installed and your network is running a dynamic host  
configuration protocol (DHCP) software package, the Ethernet option will automatically configure its  
interface with your LAN. However, it is a good idea to check these settings to make sure that the DHCP  
has successfully downloaded the appropriate network settings from your network server (See Section  
6.11.6.2).  
If your network is not running DHCP, you will have to configure the analyzer’s interface manually (See  
Section 6.11.6.3).  
26  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 360E Instruction Manual  
Getting Started  
3.3. Initial Calibration Procedure  
The next task is to calibrate the analyzer.  
To perform the following calibration you must have sources for zero air and span gas available for input into the  
sample port on the back of the analyzer. See Section 3.1.3 for instructions for connecting these gas sources.  
While it is possible to perform this procedure with any range setting we recommend that you perform this initial  
checkout using the 500 ppm range.  
NOTE  
The following procedure assumes that the instrument does not have any of the available  
Zero/Span Valve Options installed.  
See Section 7.4 for instructions for calibrating instruments possessing Z/S valve options.  
1. Set the Analog Output Range of the M360E  
SAMPLE  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
SETUP X.X  
CFG DAS RNGE PASS CLK MORE  
EXIT  
EXIT  
SETUP X.X  
RANGE CONTROL MENU  
MODE SET UNIT  
Press this button to select the  
concentration units of measure:  
Press this button to set  
the analyzer for SNGL  
DUAL or AUTO ranges  
PPB, PPM, UGM, MGM  
SETUP X.X  
RANGE: 500.000 CONC  
0
0
5
0
0
.0  
ENTR EXIT  
EXIT ignores the new setting and  
returns to the RANGE CONTROL  
MENU.  
To change the value of the  
reporting range span, enter the  
number by pressing the key under  
each digit until the expected value  
appears.  
ENTR accepts the new setting and  
SETUP X.X  
RANGE: 500.000 Conc  
.0  
returns to the  
RANGE CONTROL MENU.  
0
0
0
5
0
ENTR EXIT  
05232 Rev B3  
27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Getting Started  
Model 360E Instruction Manual  
2. Set the expected CO2 span gas concentration  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
SETUP  
< TST TST > CAL  
This sequence causes the  
analyzer to prompt for the  
expected CO2 span  
concentration.  
M-P CAL  
RANGE = 500.000 PPM  
CO2 =X.XXX  
EXIT  
< TST TST > ZERO  
CONC  
The CO2 span  
concentration values  
automatically default to  
400.0 Conc.  
EXIT ignores the new setting  
and returns to the previous  
display.  
M-P CAL  
CO2 SPAN CONC: 400.000 Conc  
To change this value to  
the actual concentration of  
the span gas, enter the  
number by pressing the  
key under each digit until  
the expected value  
ENTR accepts the new setting  
0
0
0
4
5
.0  
ENTR EXIT  
and returns to the  
previous display..  
appears.  
NOTE  
For this Initial Calibration it is important to independently verify the precise CO2  
Concentration Value of the SPAN gas.  
If the source of the Span Gas is from a Calibrated Bottle, use the exact concentration  
value printed on the bottle.  
28  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual  
Getting Started  
3. Perform the Zero/Span Calibration Procedure  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =XXX.X  
SETUP  
Set the Display to show the  
STABIL test function.  
This function calculates the  
stability of the COx  
< TST TST > CAL  
measurement  
SAMPLE  
STABIL=XXX.X PPM  
CO2 =XXX.X  
SETUP  
< TST TST > CAL  
ACTION:  
Allow zero gas to enter the sample port at the  
rear of the instrument.  
Wait until STABIL  
falls below 1.0 ppm.  
This may take several  
minutes.  
M-P CAL  
STABIL=XXX.X PPM  
CO2 =XXX.X  
< TST TST > CAL  
SETUP  
M-P CAL  
STABIL=XXX.X PPM  
CONC  
CO2 =XXX.X  
EXIT  
< TST TST > ZERO  
Press ENTR to changes the  
OFFSET & SLOPE values for the  
CO2 measurements.  
M-P CAL  
STABIL=XXX.X PPM  
CO2 =XXX.X  
< TST TST > ENTR  
CONC  
EXIT  
Press EXIT to leave the calibration  
unchanged and return to the  
previous menu.  
ACTION:  
Allow span gas to enter the sample port at the  
rear of the instrument.  
The value of  
STABIL may jump  
significantly.  
Wait until it falls back  
below 1.0 ppm  
The SPAN key now  
appears during the  
transition from zero to  
span.  
This may take several  
minutes.  
M-P CAL  
STABIL=XXX.X PPM  
CO2 =XXX.X  
EXIT  
< TST TST >  
M-P CAL  
SPAN CONC  
You may see both keys.  
If either the ZERO or  
SPAN buttons fail to  
appear see Section 11  
for troubleshooting tips.  
Press ENTR to change the  
OFFSET & SLOPE values for the  
CO2 measurements.  
RANGE = 500.000 PPM CO2 =XXX.X  
EXIT  
Press EXIT to leave the calibration  
unchanged and return to the  
previous menu.  
< TST TST > ENTR SPAN CONC  
M-P CAL  
RANGE = 500.000 PPM CO2 =XXX.X  
CONC EXIT  
EXIT returns to the main  
SAMPLE display  
< TST TST > ENTR  
05232 Rev B3  
29  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Getting Started  
Model 360E Instruction Manual  
3.3.1. Initial O2 Sensor Calibration Procedure  
If your instrument has an O2 sensor option installed that should be calibrated as well.  
3.3.1.1. O2 Calibration Setup  
The pneumatic connections for calibrating are as follows:  
Figure 3-13: O2 Sensor Calibration Set Up  
O2 SENSOR ZERO GAS: Teledyne Instruments’ recommends using pure N2 when calibration the zero point of  
your O2 sensor option.  
O2 SENSOR SPAN GAS: Teledyne Instruments’ recommends using 21% O2 in N2 when calibration the span  
point of your O2 sensor option.  
3.3.1.2. O2 Calibration Method  
STEP 1 – SET O2 SPAN GAS CONCENTRATION :  
Set the expected O2 span gas concentration.  
This should be equal to the percent concentration of the O2 span gas of the selected reporting range (default  
factory setting = 20.8%; the approximate O2 content of ambient air).  
SAMPLE  
RANGE = 500.000 PPM  
CAL  
CO2 =XXX.X  
< TST TST >  
SETUP  
SAMPLE  
GAS TO CAL:CO2  
GAS TO CAL:O2  
CO2 O2  
ENTR EXIT  
M-P CAL  
A1:NXCNC1 =100PPM  
NOX=X.XXX  
EXIT  
<TST TST> ZERO SPAN CONC  
SAMPLE  
NOX O2  
ENTR EXIT  
M-P CAL  
0
O2 SPAN CONC:20.8%  
.8  
EXIT ignores the new  
setting and returns to  
the previous display.  
2
0
0
ENTR EXIT  
ENTR accepts the new  
setting and returns to  
the previous menu.  
The OX span concentration value automatically defaults to  
20.8 %.  
If this is not the the concentration of the span gas being  
used, toggle these buttons to set the correct concentration  
of the O2 calibration gases.  
30  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Model 360E Instruction Manual  
Getting Started  
STEP 2 – ACTIVATE O2 SENSOR STABILITY FUNCTION  
To change the stability test function from NOx concentration to the O2 sensor output, press:  
SAMPLE  
RANGE = 500.000 PPM  
CO2 =XXX.X  
SETUP X.X  
0) DAS_HOLD_OFF=15.0 Minutes  
< TST TST >  
CAL  
SETUP  
<PREV NEXT> JUMP  
EDIT PRNT EXIT  
SETUP X.X  
PRIMARY SETUP MENU  
CFG DAS RNGE PASS CLK MORE  
EXIT  
EXIT  
Continue pressing NEXT until ...  
SETUP X.X  
SECONDARY SETUP MENU  
SETUP X.X  
2) STABIL_GAS=CO2  
COMM VARS DIAG ALRM  
<PREV NEXT> JUMP  
EDIT PRNT EXIT  
SETUP X.X  
ENTER PASSWORD:818  
SETUP X.X  
STABIL_GAS:CO2  
8
1
8
ENTR EXIT  
CO  
CO2 O2  
ENTR EXIT  
SETUP X.X  
STABIL_GAS:O2  
CO  
CO2 O2 ENTR EXIT  
Press EXIT 3  
times to return  
to SAMPLE  
menu  
NOTE  
Use the same procedure to reset the STB test function to CO2 when the O2 calibration  
procedure is complete.  
05232 Rev B3  
31  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Getting Started  
Model 360E Instruction Manual  
STEP 4 – O2 Zero/Span Calibration:  
To perform the zero/span calibration procedure:  
32  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual  
Getting Started  
The Model 360E Analyzer is now ready for operation  
NOTE  
Once you have completed the above set-up procedures, please fill out the Quality  
Questionnaire that was shipped with your unit and return it to Teledyne Instruments.  
This information is vital to our efforts in continuously improving our service and our  
products.  
THANK YOU.  
User Notes  
05232 Rev B3  
33  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual  
Frequently Asked Questions  
4. FREQUENTLY ASKED QUESTIONS  
4.1. FAQ’s  
The following is a list from the Teledyne Instruments’ Customer Service Department of the most commonly  
asked questions relating to the Model CO2 Analyzer.  
Q: How do I get the instrument to zero / Why is the zero key not displayed?  
A: See Section 11.3.4 Inability to zero.  
Q: How do I get the instrument to span / Why is the span key not displayed?  
A: See Section 11.3.3 Inability to span.  
Q: Why does the ENTR key sometimes disappear on the Front Panel Display?  
A: During certain types of adjustments or configuration operations, the ENTR key will disappear if you select  
a setting that is nonsensical (such as trying to set the 24-hour clock to 25:00:00) or out of the allowable range  
for that parameter (such as selecting an iDAS Holdoff period of more than 20 minutes).  
Once you adjust the setting in question to an allowable value, the ENTR key will re-appear.  
Q: Is there an optional midpoint calibration?  
A: There is an optional mid point linearity adjustment; however, midpoint adjustment is applicable only to  
applications where CO2 measurements are expected above 100 ppm. Call Teledyne Instruments’ Service  
Department for more information on this topic.  
Q: How do I make the display and data logger analog input agree?  
A: This most commonly occurs when an independent metering device is used besides the data  
logger/recorded to determine gas concentration levels while calibrating the analyzer. These disagreements  
result from the analyzer, the metering device and the data logger having slightly different ground levels.  
Both the electronic scale and offset of the analog outputs can be adjusted (see Section 6.13.4.3). Alternately,  
use the data logger itself as the metering device during calibrations procedures.  
Q: How do I perform a leak check?  
A: See Section 9.3.3.  
Q: How do I measure the sample flow?  
A: Sample flow is measured by attaching a calibrated rotameter, wet test meter, or other flow-measuring  
device to the sample inlet port when the instrument is operating. The sample flow should be 800 cm3/min  
10%. See Section 9.3.4.  
Q: How long does the IR source last?  
A: Typical lifetime is about 2-3 years.  
Q: Where is the sintered filter/sample flow control orifice?  
05232 Rev B3  
35  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Frequently Asked Questions  
Model 360E Instruction Manual  
A: These components are located inside the flow control assembly that is attached to the inlet side of the  
sample pump, see Figure 3-3. See Section 11.6.1 for instructions on disassembly and replacement.  
Q: How do I set up a SEQUENCE to run a nightly calibration check?  
A: The setup of this option is located in Section 7.6.  
Q: How do I set the analog output signal range and offset?  
A: Instructions for this can be found in Section 6.13.4 which describes analog I/O configuration.  
Q: What is the averaging time for an M360E?  
A: The default averaging time, optimized for ambient pollution monitoring, is 150 seconds for stable  
concentrations and 10 seconds for rapidly changing concentrations; See Section10.5.1 for more information.  
However, it is adjustable over a range of 0.5 second to 200 seconds (please contact customer service for  
more information).  
4.2. Glossary  
APICOM – Name of a remote control program offered by Teledyne-API to its customers  
ASSY - acronym for Assembly  
cm3 – metric abbreviation for cubic centimeter. Same as the obsolete abbreviation “cc”.  
Chemical formulas used in this document:  
CO2 – carbon dioxide  
H2O – water vapor  
O2 - molecular oxygen  
O3 - ozone  
DAS - acronym for data acquisition system, the old acronym of iDAS.  
DIAG - acronym for diagnostics, the diagnostic settings of the analyzer  
DHCP: acronym for dynamic host configuration protocol. A protocol used by LAN or Internet servers to  
automatically set up the interface protocols between themselves and any other addressable device connected to  
the network.  
DOC - Disk On Chip, the analyzer’s central storage area for analyzer firmware, configuration settings and data.  
This is a solid state device without mechanically moving parts that acts as a computer hard disk drive under  
DOS with disk label “C”. DOC chips come with 8 Mb in the E-series analyzer standard configuration but are  
available in larger sizes.  
DOS - Disk Operating System. The E-series analyzers uses DR DOS  
EEPROM - also referred to as a FLASH chip.  
FLASH - flash memory is non-volatile, solid-state memory.  
GFC – Acronym for Gas Filter Correlation.  
I2C bus - a clocked, bi-directional, serial bus for communication between individual analyzer components  
36  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 360E Instruction Manual  
Frequently Asked Questions  
IC – Acronym for Integrated Circuit, a modern, semi-conductor circuit that can contain many basic components  
such as resistors, transistors, capacitors etc in a miniaturized package used in electronic assemblies.  
iDAS - acronym for internal data acquisition system  
IP – acronym for internet protocol  
LAN - acronym for local area network  
LED - acronym for light emitting diode  
PCA - acronym for printed circuit assembly, the PCB with electronic components, ready to use.  
PCB - acronym for printed circuit board, the bare board without electronic components  
PLC – Acronym for programmable logic controller, a device that is used to control instruments based on a logic  
level signal coming from the analyzer  
PFA – Acronym for Per-Fluoro-Alkoxy, an inert polymer. One of the polymers that du Pont markets as Teflon®  
(along with FEP and PTFE).  
PTFE – Acronym for Poly-Tetra-Fluoro-Ethylene, a very inert polymer material used to handle gases that may  
react on other surfaces. One of the polymers that du Pont markets as Teflon® (along with FEP and PFA).  
PVC – Acronym for Poly Vinyl Chloride, a polymer used for downstream tubing in the M360E.  
RS-232 - an electronic communications type of a serial communications port  
RS-485 - an electronic communications type of a serial communications port  
TCP/IP - acronym for transfer control protocol / internet protocol, the standard communications protocol for  
Ethernet devices.  
VARS - acronym for variables, the variables settings of the analyzer  
User Notes  
05232 Rev B3  
37  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual  
Optional Hardware and Software  
5. OPTIONAL HARDWARE AND SOFTWARE  
This includes a brief description of the hardware and software options available for the Model 360E Gas Filter  
Correlation Carbon Dioxide Analyzer. For assistance with ordering these options please contact the Sales  
department of Teledyne – Advanced Pollution Instruments at:  
TOLL-FREE: 800-324-5190  
FAX: 858-657-9816  
TEL: 858-657-9800  
E-MAIL: apisales@teledyne.com  
WEB SITE: www.teledyne-api.com  
5.1. Rack Mount Kits (Options 20a, 20b & 21)  
OPTION NUMBER  
OPT 20A  
DESCRIPTION  
Rack mount brackets with 26 in. chassis slides.  
Rack mount brackets with 24 in. chassis slides.  
Rack mount brackets only  
OPT 20B  
OPT 21  
Each of these options permits the Analyzer to be mounted in a standard 19" x 30" RETMA rack.  
5.2. Current Loop Analog Outputs (Option 41)  
This option adds isolated, voltage-to-current conversion circuitry to the analyzer’s analog outputs. This option  
may be ordered separately for any of the analog outputs; it can be installed at the factory or added later. Call T-  
API sales for pricing and availability.  
The current loop option can be configured for any output range between 0 and 20 mA. Information on calibrating  
or adjusting these outputs can be found in Section 6.13.4.4.  
J19  
J 23  
Voltage Output  
Shunts installed  
Voltage Output  
Shunts installed  
Current Loop Option  
Installed on J21  
(Analog Output A2)  
Figure 5-1: Current Loop Option Installed on the Motherboard  
05232 Rev B3  
39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Optional Hardware and Software  
Model 360E Instruction Manual  
5.2.1. Converting Current Loop Analog Outputs to Standard Voltage  
Outputs.  
NOTE  
Servicing or handling of circuit components requires electrostatic discharge protection,  
i.e. ESD grounding straps, mats and containers. Failure to use ESD protection when  
working with electronic assemblies will void the instrument warranty.  
See Chapter 12 for more information on preventing ESD damage.  
To convert an output configured for current loop operation to the standard 0 to 5 VDC output operation:  
4. Turn off power to the analyzer.  
5. If a recording device was connected to the output being modified, disconnect it.  
6. Remove the top cover  
Remove the set screw located in the top, center of the rear panel  
Remove the screws fastening the top cover to the unit (four per side).  
Lift the cover straight up.  
7. Disconnect the current loop option PCA from the appropriate connector on the motherboard (see Figure  
5-1).  
8. Place a shunt between the leftmost two pins of the connector (see Figure 5-1).  
6 spare shunts (P/N CN0000132) were shipped with the instrument attached to JP1 on the back of the  
instruments keyboard and display PCA  
9. Reattach the top case to the analyzer.  
10. The analyzer is now ready to have a voltage-sensing, recording device attached to that output  
5.3. Expendable Kits (Options 42C, 42D and 43)  
OPTION NUMBER  
OPT 42C  
DESCRIPTION  
1 year’s supply of replacement of 47mm dia. particulate filters  
OPT 42D  
1 full replacement’s volume of indicating soda-lime for the external CO2 scrubber  
included with options 51 & 53 (approximate active lifetime: 1 year)  
OPT 43  
Options 42 C & 42D  
40  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Model 360E Instruction Manual  
Optional Hardware and Software  
5.4. Calibration Valves Options  
There are four available options involving Zero/Span/Shutoff valves. From an operational and software  
standpoint, all of the options are the same, only the source of the span and zero gases are different.  
5.4.1. Zero/Span/Shutoff Valve (Option 50)  
This option requires that both zero air and span gas be supplied from external sources. It is specifically  
designed for applications where span gas will be supplied from a pressurized bottle of calibrated CO2 gas. A  
critical flow control orifice, internal to the instrument ensures that the proper flow rate is maintained. An internal  
vent line, isolated by a shutoff valve ensures that the gas pressure of the span gas is reduced to ambient  
atmospheric pressure. Normally zero air would be supplied from zero air modules such as a Teledyne  
Instruments Model 701.  
In order to ensure that span gas does not migrate backwards through the vent line and alter the concentration of  
the span gas, a gas line not less than 2 meters in length should be attached to the vent span outlet on the rear  
panel of the analyzer. To prevent the buildup of back pressure, this vent line should not be greater than 10  
meters in length.  
The following table describes the state of each valve during the analyzer’s various operational modes.  
Table 5-1: Zero/Span Valve Operating States for Options 50 & 51  
MODE  
VALVE  
CONDITION  
Sample/Cal  
Zero/Span  
Open to SAMPLE inlet  
Open to IZS inlet  
Closed  
SAMPLE  
(Normal State)  
Shutoff Valve  
Sample/Cal  
Zero/Span  
Open to ZERO/SPAN valve  
Open to IZS inlet  
Closed  
ZERO CAL  
SPAN CAL  
Shutoff Valve  
Sample/Cal  
Zero/Span  
Open to ZERO/SPAN valve  
Open to SHUTOFF valve  
Shutoff Valve  
Open to PRESSURE SPAN Inlet  
The minimum span gas flow rate required for this option is 800 cm3/min.  
The state of the zero/span valves can also be controlled:  
Manually from the analyzer’s front panel by using the SIGNAL I/O controls located under the DIAG  
Menu (Section 6.13.2),  
By activating the instrument’s AutoCal feature (Section 7.6),  
Remotely by using the external digital control inputs (Section 6.15.1.2 and Section 7.5.2), or;  
Remotely through the RS-232/485 serial I/O ports (see Appendix A-6 for the appropriate commands).  
05232 Rev B3  
41  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Optional Hardware and Software  
Model 360E Instruction Manual  
Figure 5-2: Internal Pneumatic Flow – Zero/Span/Shutoff Valves OPT 50 & 51  
5.4.2. Zero/Span/Shutoff with External CO2 Scrubber  
(Option 51)  
Option 51 is operationally and pneumatically identical to Option 50 above (See Section5.4.1), except that the  
zero air is generated by an externally mounted zero air scrubber filled with indicating soda-lime that changes  
color from white to pink as it becomes saturated.  
5.4.3. Zero/Span Valve (Option 52)  
This valve option is intended for applications where zero air is supplied by a zero air generator like the Teledyne  
Instruments Model 701 and span gas are being supplied by Gas Dilution Calibrator like the Teledyne  
Instruments Model 700 or 702. Internal zero/span and sample/cal valves control the flow of gas through the  
instrument, but because the calibrator limits the flow of span gas no shutoff valve is required.  
In order to ensure that span gas does not migrate backwards through the vent line and alter the concentration of  
the span gas, a gas line not less than 2 meters in length should be attached to the vent span outlet on the rear  
panel of the analyzer. To prevent the buildup of back pressure, this vent line should not be greater than 10  
meters in length.  
The following table describes the state of each valve during the analyzer’s various operational modes.  
42  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Model 360E Instruction Manual  
Optional Hardware and Software  
Table 5-2: Zero/Span Valve Operating States for Options 52 & 53  
Mode  
Valve  
Condition  
Sample/Cal  
Zero/Span  
Open to SAMPLE inlet  
Open to IZS inlet  
SAMPLE  
(Normal State)  
Sample/Cal  
Zero/Span  
Open to ZERO/SPAN valve  
Open to IZS inlet  
ZERO CAL  
SPAN CAL  
Sample/Cal  
Zero/Span  
Open to ZERO/SPAN valve  
Open to PRESSURE SPAN inlet  
The minimum span gas flow rate required for this option is 800 cm3/min.  
The state of the zero/span valves can also be controlled:  
Manually from the analyzer’s front panel by using the SIGNAL I/O controls located under the DIAG  
Menu (Section 6.13.2),  
By activating the instrument’s AutoCal feature (Section 7.6),  
Remotely by using the external digital control inputs (Sections 6.15.1.2 and 7.5.2), or  
Remotely through the RS-232/485 serial I/O ports (see Appendix A-6).  
Figure 5-3: Internal Pneumatic Flow – Zero/Span OPT 52 & 53  
05232 Rev B3  
43  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Optional Hardware and Software  
Model 360E Instruction Manual  
5.4.4. Zero/Span Valve with External CO2 Scrubber (Option 53)  
Option 53 is operationally and pneumatically identical to Option 52 above (See Section5.4.3), except that the  
zero air is generated by an externally mounted zero air scrubber filled with indicating soda-lime that changes  
color from white to pink as it becomes saturated.  
5.5. Communication Options  
5.5.1. RS232 Modem Cable (Option 60)  
The analyzer can have come standard with a shielded, straight-through DB-9F to DB-9F cable of about 1.8 m  
length, which should fit most computers of recent build. This cable can be ordered as Option 60.  
Option 60A consists of a shielded, straight-through serial cable of about 1.8 m length to connect the analyzer’s  
COM1 port to a computer, a code activated switch or any other communications device that is equipped with a  
DB-25 female connector. The cable is terminated with one DB-9 female connector and one DB-25 male  
connector. The DB-9 connector fits the analyzer’s COM1 port.  
Some older computers or code activated switches with a DB-25 serial connector will need a different cable or an  
appropriate adapter.  
5.5.2. RS-232 Multidrop (Option 62)  
The multidrop option is used with any of the RS-232 serial ports to enable communications of up to eight  
analyzers with the host computer over a chain of RS-232 cables via the instruments COM1 Port. It is subject to  
the distance limitations of the RS 232 standard.  
The option consists of a small printed circuit assembly, which is plugs into to the analyzer’s CPU card (see  
Figure 5-4) and is connected to the RS-232 and COM2 DB9 connectors on the instrument’s back panel via a  
cable to the motherboard. One option 62 is required for each analyzer along with one 6’ straight-through, DB9  
male DB9 Female cable (P/N WR0000101).  
This option can be installed in conjunction with the Ethernet option (Option 63) allowing the instrument to  
communicate on both types of networks simultaneously. For more information on using and setting up this  
option (See Section 6.11.7)  
CPU Card  
Rear Panel  
(as seen from inside)  
Multidrop  
Card  
Figure 5-4: M360E Multidrop Card  
44  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Model 360E Instruction Manual  
Optional Hardware and Software  
5.5.3. Ethernet (Option 63)  
When installed, this option is electronically connected to the instrument’s COM2 serial port making that port no  
longer available for RS-232/RS-485 communications through the COM2 connector on the rear panel. The  
option consists of a Teledyne Instruments designed Ethernet card (see Figure 5-5), which is mechanically  
attached to the instrument’s rear panel (see Figure 5-6). A 7-foot long CAT-5 network cable, terminated at both  
ends with standard RJ-45 connectors, is included as well. Maximum communication speed is limited by the RS-  
232 port to 115.2 kBaud.  
Figure 5-5: M360E Ethernet Card  
CPU  
Card  
Rear Panel  
(as seen from inside)  
Ethernet  
Card  
Female RJ-45  
Connector  
LNK LED  
ACT LED  
TxD LED  
RxD LED  
RE-232  
Connector To  
Motherboard  
Interior View  
Exterior View  
Figure 5-6: M360E Rear Panel with Ethernet Installed  
This option can be installed in conjunction with the RS-2323 multidrop (option 62) allowing the instrument to  
communicate on both types of networks simultaneously. For more information on using and setting up this  
option. See Section 6.11.6)  
05232 Rev B3  
45  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Optional Hardware and Software  
Model 360E Instruction Manual  
5.6. Oxygen Sensor (OPT 65)  
5.6.1. Theory of Operation  
5.6.1.1. Paramagnetic measurement of O2  
The oxygen sensor used in the M360E analyzer utilizes the fact that oxygen is attracted into strong magnetic  
field; most other gases are not, to obtain fast, accurate oxygen measurements.  
The sensor’s core is made up of two nitrogen filled glass spheres, which are mounted on a rotating suspension  
within a magnetic field (Figure 5-7). A mirror is mounted centrally on the suspension and light is shone onto the  
mirror that reflects the light onto a pair of photocells. The signal generated by the photocells is passed to a  
feedback loop, which outputs a current to a wire winding (in effect, a small DC electric motor) mounted on the  
suspended mirror.  
Oxygen from the sample stream is attracted into the magnetic field displacing the nitrogen filled spheres and  
causing the suspended mirror to rotate. This changes the amount of light reflected onto the photocells and  
therefore the output levels of the photocells. The feedback loop increases the amount of current fed into the  
winding in order to move the mirror back into its original position. The more O2 present, the more the mirror  
moves and the more current is fed into the winding by the feedback control loop.  
A sensor measures the amount of current generated by the feedback control loop which is directly proportional  
to the concentration of oxygen within the sample gas mixture (see Figure 5-7).  
Figure 5-2: Oxygen Sensor - Principle of Operation  
5.6.1.2. Operation within the M360E Analyzer  
The oxygen sensor option is transparently integrated into the core analyzer operation. All functions can be  
viewed or accessed through the front panel, just like the functions for CO2  
The O2 concentration is displayed in the upper right-hand corner, alternating with CO2 concentration.  
Test functions for O2 slope and offset are viewable from the front panel along with the analyzer’s other  
test functions.  
O2 sensor calibration is performed via the front panel CAL function and is performed in a nearly identical  
manner as the standard CO2 calibration. See Section 3.3.1 for more details.  
Stability of the O2 sensor can be viewed via the front panel (see Section 3.3.2.1).  
A signal representing the currently measured O2 concentration is available.  
46  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Model 360E Instruction Manual  
Optional Hardware and Software  
The O2 concentration range is 0-100% (user selectable) with 0.1% precision and accuracy and is available to be  
output via the instrument’s analog output channel A3 (See Section 6.13.4).  
The temperature of the O2 sensor is maintained at a constant 50° C by means of a PID loop and can be viewed  
on the front panel as test function O2 TEMP.  
The O2 sensor assembly itself does not have any serviceable parts and is enclosed in an insulated canister.  
5.6.1.3. Pneumatic Operation of the O2 Sensor  
Pneumatically, the O2 sensor is connected to the bypass manifold and draws a flow of about 120  
cm³/min in addition to the normal sample flow rate and is separately controlled with its own critical flow  
orifice. Figure 5-8 shows the internal pneumatics of the M360E with the O2 Sensor installed.  
Figure 5-3: M360E – Internal Pneumatics with O2 Sensor Option 65  
05232 Rev B3  
47  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Optional Hardware and Software  
Model 360E Instruction Manual  
5.7. Additional Manuals  
5.7.1. Printed Manuals (Option 70)  
Additional printed copies of this manual are available from Teledyne Instruments  
5.7.2. Manual on CD (Part number 045840200)  
This operator’s manual is also available on CD. The electronic document is stored in Adobe Systems Inc.  
Portable Document Format (PDF) and is viewable with Adobe Acrobat Reader® software, downloadable for free  
at http://www.adobe.com/  
The CD version of the manual has many advantages:  
Fully searchable text.  
Hypertext links for figures, tables, table of contents and embedded references for quick access of  
individual manual portions.  
A list of thumbnails, chapters and s displayed at the left of the text.  
Internet links embedded in the manual will take you to the corresponding web site (requires an internet  
connection).  
5.8. Extended Warranty (Options 92 & 93)  
Two options are available for extending Teledyne Instruments’ standard warranty (Section 2.3). Both options  
have to be specified upon ordering the analyzer.  
OPTION  
DESCRIPTION  
NUMBER  
OPT 92  
OPT 93  
Extends warranty to cover a two (2) year period from the date of purchase.  
Extends warranty to cover a five (5) year period from the date of purchase.  
5.9. Special Software Features  
5.9.1. Dilution Ratio Option  
The Dilution Ration Option is a software option that is designed for applications where the Sample gas is diluted  
before being analyzed by the Model 360E. Typically this occurs in Continuous Emission Monitoring (CEM)  
applications where the quality of gas in a smoke stack is being tested and the sampling method used to remove  
the gas from the stack dilutes the gas.  
Once the degree of dilution is known, this feature allows the user to add an appropriate scaling factor to the  
analyzer’s CO2 concentration calculation so that the Measurement Range and concentration values displayed on  
the instrument’s Front Panel Display and reported via the Analog and Serial Outputs reflect the undiluted values.  
Instructions for using the dilution ratio option can be found in Section 6.8.7.  
48  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
           
Model 360E Instruction Manual  
Optional Hardware and Software  
5.9.2. Maintenance Mode Switch  
API’s instruments can be equipped with a switch that places the instrument in maintenance mode. When  
present, the switch accessed by opening the hinged front panel and is located on the rearward facing side of the  
display/keyboard driver PCA; on the left side; near the particulate filter.  
When in maintenance mode the instrument ignores all commands received via the COMM ports that alter the  
operation state of the instrument. This includes all calibration commands, diagnostic menu commands and the  
reset instrument command. The instrument continues to measure concentration and send data when requested.  
This option is of particular use for instruments connected to multidrop or Hessen protocol networks.  
5.9.3. Second Language Switch  
API’s instruments can be equipped with switch that activates an alternate set of display message in a language  
other than the instruments default language. When present, the switch accessed by opening the hinged front  
panel and is located on the rearward facing side of the display/keyboard driver PCA; on the right side.  
To activate this feature, the instrument must also have a specially programmed Disk on Chip containing the  
second language.  
User Notes  
05232 Rev B3  
49  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual  
Operating Instructions  
6. OPERATING INSTRUCTIONS  
To assist in navigating the analyzer’s software, a series of menu trees can be found in Appendix A-1 of this  
manual.  
NOTES  
The flow charts appearing in this contain typical representations of the analyzer’s  
display during the various operations being described. These representations may  
differ slightly from the actual display of your instrument.  
The ENTR key may disappear if you select a setting that is invalid or out of the allowable  
range for that parameter, such as trying to set the 24-hour clock to 25:00:00. Once you  
adjust the setting to an allowable value, the ENTR key will re-appear.  
6.1. Overview of Operating modes  
The M360E software has a variety of operating modes. Most commonly, the analyzer will be operating in  
SAMPLE mode. In this mode, a continuous read-out of the CO2 concentration is displayed on the front panel  
and output as an analog voltage from rear panel terminals, calibrations can be performed, and TEST functions  
and WARNING messages can be examined.  
The second most important operating mode is SETUP mode. This mode is used for performing certain  
configuration operations, such as for the iDAS system, the reporting ranges, or the serial (RS-232/RS-  
485/Ethernet) communication channels. The SET UP mode is also used for performing various diagnostic tests  
during troubleshooting.  
Mode Field  
SAMPLE A  
RANGE = 500.00 PPM  
CO2 400.00  
SETUP  
<TST TST> CAL  
Figure 6-1: Front Panel Display  
The mode field of the front panel display indicates to the user which operating mode the unit is currently running.  
Besides SAMPLE and SETUP, other modes the analyzer can be operated in are:  
05232 Rev B3  
51  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Operating Instructions  
Model 360E Instruction Manual  
Table 6-1: Analyzer Operating modes  
MEANING  
MODE  
DIAG  
One of the analyzer’s diagnostic modes is being utilized (See Section 6.13).  
M-P CAL  
This is the basic, multi-point calibration mode of the instrument and is activated by pressing  
the CAL key.  
SAMPLE  
SAMPLE A  
SETUP1  
Sampling normally, flashing indicates adaptive filter is on.  
Indicates that unit is in SAMPLE Mode and AUTOCAL feature is activated.  
SETUP mode is being used to configure the analyzer (CO2 sampling will continue during this  
process).  
SPAN CAL A  
Unit is performing span cal procedure initiated automatically by the analyzer’s AUTOCAL  
feature.  
SPAN CAL M  
SPAN CAL R  
Unit is performing span cal procedure initiated manually by the user.  
Unit is performing span cal procedure initiated remotely via the RS-232, RS-4485 or digital i/o  
control inputs.  
ZERO CAL A  
Unit is performing zero cal procedure initiated automatically by the analyzer’s AUTOCAL  
feature.  
ZERO CAL M  
ZERO CAL R  
Unit is performing zero cal procedure initiated manually by the user.  
Unit is performing zero cal procedure initiated remotely via the RS-232, RS-4485 or digital I/O  
control inputs.  
1 The revision of the Teledyne Instruments software installed in this analyzer will be displayed following the word  
SETUP. E.g. “SETUP G.4”  
Finally, the various CAL modes allow calibration of the analyzer. Because of its importance, this mode is  
described separately in Chapter 7.  
6.2. Sample Mode  
This is the analyzer’s standard operating mode. In this mode the instrument is analyzing the gas in the sample  
chamber, calculating CO2 concentration and reporting this information to the user via the front panel display, the  
analog outputs and, if set up properly, the RS-232/485/Ethernet ports.  
NOTE  
A value of “XXXX” displayed in the CO2 Concentration field means that the M/R ratio is  
invalid because CO2 REF is either too high (> 4950 mVDC) or too low (< 1250 VDC).  
6.2.1. Test Functions  
A series of test functions is available at the front panel while the analyzer is in SAMPLE mode. These  
parameters provide information about the present operating status of the instrument and are useful during  
troubleshooting (Section 11.1.2). They can also be recorded in one of the iDAS channels (Section 6.7) for data  
analysis. To view the test functions, press one of the <TST TST> keys repeatedly in either direction.  
52  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Model 360E Instruction Manual  
Operating Instructions  
Table 6-2: Test Functions Defined  
UNITS  
DISPLAY  
PARAMETER  
MEANING  
TITLE  
Range  
RANGE  
The full scale limit at which the reporting range of the analyzer is  
currently set. THIS IS NOT the Physical Range of the instrument.  
See Section 6.8.1 for more information.  
PPB, PPM,  
UGM, MGM  
RANGE11  
RANGE21  
O2 Range  
Stability  
The range setting for the O2 Sensor  
O2 RANGE  
STABIL  
%
PPB, PPM  
UGM, MGM  
Standard deviation of CO2 concentration readings. Data points are  
recorded every ten seconds using the last 25 data points.  
CO2 Measure  
The demodulated, peak IR detector output during the measure  
portion of the CFG Wheel cycle.  
MEAS  
REF  
MV  
MV  
CO2 Reference  
The demodulated, peak IR detector output during the reference  
portion of the CFG wheel cycle.  
Measurement /  
Reference Ratio  
The result of CO2 MEAS divided by CO2 REF. This ratio is the  
primary value used to compute CO2 concentration. The value  
displayed is not linearized.  
MR Ratio  
PRES  
-
Sample Pressure  
Sample Flow  
The absolute pressure of the Sample gas as measured by a  
pressure sensor located inside the sample chamber.  
In-Hg-A  
cc/min  
Sample mass flow rate. This is computed from the differential  
between the pressures measured up-stream and down-stream of the  
sample critical flow orifice pressures.  
SAMPLE FL  
SAMP TEMP  
Sample  
Temperature  
The temperature of the gas inside the sample chamber.  
Optical bench temperature.  
C  
Bench  
Temperature  
BENCH  
TEMP  
C  
Wheel  
Temperature  
WHEEL  
TEMP  
Filter wheel temperature.  
C  
C  
C  
Box Temperature  
The temperature inside the analyzer chassis.  
O2 sensor cell temperature.  
BOX TEMP  
O2 Cell  
O2 CELL  
TEMP2  
Temperature2  
Photo-detector  
Temp. Control  
Voltage  
The drive voltage being supplied to the thermoelectric coolers of the  
IR photo-detector by the sync/demod Board.  
PHT DRIVE  
SLOPE  
mV  
Slope  
The sensitivity of the instrument as calculated during the last  
calibration activity. The SLOPE parameter is used to set the span  
calibration point of the analyzer.  
-
-
Offset  
The overall offset of the instrument as calculated during the last  
calibration activity. The OFFSET parameter is used to set the zero  
point of the analyzer response.  
OFFSET  
O2 Sensor Slope  
O2 SLOPE  
O2 OFFSET  
TEST  
O2 slope, computed during zero/span calibration.  
O2 offset, computed during zero/span calibration.  
-
2
O2 Sensor Offset  
-
2
Test channel  
output signal  
Displays the signal level of the TEST analog output channel. Only  
appears when the TEST channel has been activated.  
mV, mA  
Current Time  
The current time. This is used to create a time stamp on iDAS  
readings, and by the AUTOCAL feature to trigger calibration events.  
TIME  
-
1 Only appears when the instrument’s reporting range mode is set for DUAL or AUTO  
2 Only appears when the optional O2 sensor is installed.  
To view the TEST Functions press the following Key sequence:  
05232 Rev B3  
53  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Operating Instructions  
Model 360E Instruction Manual  
SAMPLE  
RANGE = 500.000 PPM  
CO2 = XXX.X  
SETUP  
< TST TST > CAL  
RANGE  
RANGE11  
RANGE21  
02 RANGE  
STABIL  
Toggle <TST TST> keys to  
scroll through list of functions  
MEAS  
REF  
MR RATIO  
PRES  
Refer to  
Table 6-2 for  
definitions of  
these test  
SAMP FL  
SAMP TEMP  
BENCH TEMP  
WHEEL TEMP  
BOX TEMP  
PHT DRIVE  
SLOPE  
functions.  
1 Only appears instrument is set  
for DUAL or AUTO reporting  
range modes.  
OFFSET  
TEST  
TIME  
2 Only appears if 02 Sensor  
Option is installed.  
Figure 6-3: Viewing M360E TEST Functions  
NOTE  
A value of “XXXX” displayed for any of the TEST functions indicates an out-of-range  
reading or the analyzer’s inability to calculate it.  
All pressure measurements are represented in terms of absolute pressure. Absolute,  
atmospheric pressure is 29.92 in-Hg-A at sea level. It decreases about 1 in-Hg per 300  
m gain in altitude. A variety of factors such as air conditioning and passing storms can  
cause changes in the absolute atmospheric pressure.  
54  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 360E Instruction Manual  
Operating Instructions  
6.2.2. Warning Messages  
The most common instrument failures will be reported as a warning on the analyzer’s front panel and through  
the COM ports. Section 11.1.1 explains how to use these messages to troubleshoot problems. Section 3.2.3  
shows how to view and clear warning messages.  
Table 6-3 lists all warning messages for the current version of software.  
Table 6-3: List of Warning Messages  
MESSAGE  
MEANING  
The instrument’s A/D circuitry or one of its analog outputs is not calibrated.  
The Temperature of the optical bench is outside the specified limits.  
Remote span calibration failed while the dynamic span feature was set to turned on  
Remote zero calibration failed while the dynamic zero feature was set to turned on  
Configuration was reset to factory defaults or was erased.  
Concentration alarm 1 is enabled and the measured CO2 level is the set point.  
Concentration alarm 2 is enabled and the measured CO2 level is the set point.  
Configuration storage was reset to factory configuration or erased.  
iDAS data storage was erased.  
ANALOG CAL WARNING  
BENCH TEMP WARNING  
BOX TEMP WARNING  
CANNOT DYN SPAN  
CANNOT DYN ZERO  
CONC ALRM1 WARNING  
CONC ALRM2 WARNING  
CONFIG INITIALIZED  
DATA INITIALIZED  
O2 sensor cell temperature outside of warning limits.  
O2 CELL TEMP WARN  
PHOTO TEMP WARNING  
REAR BOARD NOT DET  
RELAY BOARD WARN  
SAMPLE FLOW WARN  
SAMPLE PRESS WARN  
SAMPLE TEMP WARN  
SOURCE WARNING  
The temperature of the IR photometer is outside the specified limits.  
The CPU is unable to communicate with the motherboard.  
The firmware is unable to communicate with the relay board.  
The flow rate of the sample gas is outside the specified limits.  
Sample gas pressure outside of operational parameters.  
The temperature of the sample gas is outside the specified limits.  
The IR source may be faulty.  
The computer was rebooted.  
SYSTEM RESET  
The Gas Filter Correlation wheel temperature is outside the specified limits.  
WHEEL TEMP WARNING  
05232 Rev B3  
55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Operating Instructions  
Model 360E Instruction Manual  
To view and clear warning messages  
SAMPLE  
HVPS WARNING  
CAL MSG  
CO2 = 0.00  
TEST deactivates warning  
TEST  
CLR SETUP  
messages  
MSG activates warning  
SAMPLE  
RANGE=500.000 PPM  
MSG  
CO2 = 0.00  
messages.  
<TST TST> keys replaced with  
< TST TST > CAL  
CLR SETUP  
TEST key  
SAMPLE  
HVPS WARNING  
CO2 = 0.00  
Press CLR to clear the current  
message.  
TEST  
CAL  
MSG  
CLR SETUP  
NOTE:  
If more than one warning is active, the  
next message will take its place  
If the warning message persists  
after several attempts to clear it,  
the message may indicate a  
real problem and not an artifact  
of the warm-up period  
Once the last warning has been  
cleared, the analyzer returns to  
SAMPLE mode  
Make sure warning messages are  
not due to real problems.  
Figure 6-3  
Viewing and Clearing M360E WARNING Messages  
6.3. Calibration Mode  
Pressing the CAL key switches the M360E into multi-point calibration mode. In this mode, the user can calibrate  
the instrument or check the instruments calibration with the use of calibrated zero or span gases.  
If the instrument includes either the zero/span valve option or IZS option, the display will also include CALZ and  
CALS keys. Pressing either of these keys also puts the instrument into multipoint calibration mode.  
The CALZ key is used to initiate a calibration of the zero point.  
The CALS key is used to calibrate the span point of the analyzer. It is recommended that this span  
calibration is performed at 90% of full scale of the analyzer’s currently selected reporting range.  
Because of their critical importance and complexity, calibration operations are described in detail in Chapter 7 of  
this manual. For more information concerning the zero/span, zero/span/shutoff and IZS valve options, See  
Section 5.4.  
56  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Model 360E Instruction Manual  
Operating Instructions  
6.4. SETUP MODE  
The SETUP mode contains a variety of choices that are used to configure the analyzer’s hardware and software  
features, perform diagnostic procedures, gather information on the instruments performance and configure or  
access data from the internal data acquisition system (iDAS). For a visual representation of the software menu  
trees, refer to Appendix A-1.  
The areas access under the Setup mode are:  
Table 6-4: Primary Setup Mode Features and Functions  
KEYPAD  
LABEL  
MANUAL  
SECTION  
MODE OR FEATURE  
DESCRIPTION  
Analyzer Configuration  
CFG  
Lists key hardware and software configuration information  
Used to set up and operate the AutoCal feature.  
6.5  
Auto Cal Feature  
ACAL  
7.6  
Only appears if the analyzer has one of the internal valve  
options installed  
Internal Data Acquisition  
(iDAS)  
DAS  
Used to set up the iDAS system and view recorded data  
6.7  
6.8  
Analog Output Reporting  
Range Configuration  
Used to configure the output signals generated by the  
instruments Analog outputs.  
RNGE  
Calibration Password Security  
Internal Clock Configuration  
PASS  
CLK  
Turns the calibration password feature ON/OFF  
Used to Set or adjust the instrument’s internal clock  
6.9  
6.10  
See  
Table 6-5  
Advanced SETUP features  
MORE  
This button accesses the instruments secondary setup menu  
Table 6-5: Secondary Setup Mode Features and Functions  
KEYPAD  
MANUAL  
SECTION  
MODE OR FEATURE  
DESCRIPTION  
LABEL  
Used to set up and operate the analyzer’s various external I/O  
channels including RS-232; RS 485, modem communication  
and/or Ethernet access.  
External Communication  
Channel Configuration  
6.11 &  
6.15  
COMM  
VARS  
DIAG  
Used to view various variables related to the instruments current  
operational status  
System Status Variables  
6.12  
6.13  
Used to access a variety of functions that are used to configure,  
test or diagnose problems with a variety of the analyzer’s basic  
systems  
System Diagnostic Features  
Used to activate the analyzer’s two gas concentration status  
alarms and set the alarm limits  
CO2 Concentration Alarms  
ALRM  
6.14  
NOTE  
Any changes made to a variable during one of the following procedures is not  
acknowledged by the instrument until the ENTR Key is pressed  
If the EXIT key is pressed before the ENTR key, the analyzer will beep alerting the user  
that the newly entered value has been lost.  
05232 Rev B3  
57  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Operating Instructions  
Model 360E Instruction Manual  
6.5. SETUP CFG: Viewing the Analyzer’s Configuration  
Information  
Pressing the CFG key displays the instrument configuration information. This display lists the analyzer model,  
serial number, firmware revision, software library revision, CPU type and other information. Use this information  
to identify the software and hardware when contacting customer service. Special instrument or software  
features or installed options may also be listed here.  
SAMPLE*  
RANGE = 500.000 PPB  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
Press NEXT of PREV to move back and  
forth through the following list of  
Configuration information:  
MODEL NAME  
SAMPLE  
PRIMARY SETUP MENU  
Press EXIT at any  
time to return to the  
SAMPLE display  
SERIAL NUMBER  
SOFTWARE REVISION  
CFG DAS RNGE PASS CLK MORE  
EXIT  
EXIT  
LIBRARY REVISION  
iCHIP SOFTWARE REVISION1  
HESSEN PROTOCOL REVISION1  
ACTIVE SPECIAL SOFTWARE  
OPTIONS1  
SAMPLE  
M360E CO2 ANALYZER  
Press EXIT at  
any time to  
return to SETUP  
menu  
CPU TYPE  
DATE FACTORY CONFIGURATION  
SAVED  
NEXT PREV  
1Only appears if relevant option of Feature is active.  
6.6. SETUP ACAL: Automatic Calibration  
Instruments with one of the internal valve options installed can be set to automatically run calibration procedures  
and calibration checks. These automatic procedures are programmed using the submenus and functions found  
under the ACAL menu.  
A menu tree showing the ACAL menu’s entire structure can be found in Appendix A-1 of this manual.  
Instructions for using the ACAL feature are located in the Section 7.6 of this manual along with all other  
information related to calibrating the M360E analyzer.  
58  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Model 360E Instruction Manual  
Operating Instructions  
6.7. SETUP DAS: Using the Data Acquisition System (iDAS)  
The M360E analyzer contains a flexible and powerful, internal data acquisition system (iDAS) that enables the  
analyzer to store concentration and calibration data as well as a host of diagnostic parameters. The iDAS of the  
M360E can store up to about one million data points, which can, depending on individual configurations, cover  
days, weeks or months of valuable measurements. The data are stored in non-volatile memory and are retained  
even when the instrument is powered off. Data are stored in plain text format for easy retrieval and use in  
common data analysis programs (such as spreadsheet-type programs).  
The iDAS is designed to be flexible, users have full control over the type, length and reporting time of the data.  
The iDAS permits users to access stored data through the instrument’s front panel or its communication ports.  
Using APICOM, data can even be retrieved automatically to a remote computer for further processing.  
The principal use of the iDAS is logging data for trend analysis and predictive diagnostics, which can assist in  
identifying possible problems before they affect the functionality of the analyzer. The secondary use is for data  
analysis, documentation and archival in electronic format.  
To support the iDAS functionality, Teledyne Instruments offers APICOM, a program that provides a visual  
interface for remote or local setup, configuration and data retrieval of the iDAS (Section 6.7). The APICOM  
manual, which is included with the program, contains a more detailed description of the iDAS structure and  
configuration.  
The M360E is configured with a basic iDAS configuration, which is enabled by default. New data channels are  
also enabled by default but each channel may be turned off for later or occasional use. Note that iDAS  
operation is suspended while its configuration is edited through the front panel. To prevent such data loss, it is  
recommended to use the APICOM graphical user interface for iDAS changes.  
The green SAMPLE LED on the instrument front panel, which indicates the analyzer status, also indicates  
certain aspects of the iDAS status:  
Table 6-6: Front Panel LED Status Indicators for iDAS  
LED STATE  
OFF  
iDAS Status  
System is in calibration mode. Data logging can be enabled or disabled for this mode.  
Calibration data are typically stored at the end of calibration periods, concentration data are  
typically not sampled, diagnostic data should be collected.  
BLINKING  
ON  
Instrument is in hold-off mode, a short period after the system exits calibrations. IDAS  
channels can be enabled or disabled for this period. Concentration data are typically disabled  
whereas diagnostic should be collected.  
Sampling normally.  
The iDAS can be disabled only by disabling or deleting its individual data channels.  
6.7.1. iDAS Structure  
The iDAS is designed around the feature of a “record”. A record is a single data point of one parameter, stored  
in one (or more) data channels and generated by one of several triggering event. The entire iDAS configuration  
is stored in a script, which can be edited from the front panel or downloaded, edited and uploaded to the  
instrument in form of a string of plain-text lines through the communication ports.  
iDAS data are defined by the PARAMETER type and are stored through different triggering EVENTS in data  
CHANNELS, which relate triggering events to data parameters and define certain operational functions related  
to the recording and reporting of the data.  
05232 Rev B3  
59  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Operating Instructions  
Model 360E Instruction Manual  
6.7.1.1. iDAS Channels  
The key to the flexibility of the iDAS is its ability to store a large number of combinations of triggering events and  
data parameters in the form of data channels. Users may create up to 20 data channels and each channel can  
contain one or more parameters. For each channel one triggering event is selected and up to 50 data  
parameters, which can be the same or different between channels. Each data channel has several properties  
that define the structure of the channel and allow the user to make operational decisions regarding the channel  
(Table 6-7).  
Table 6-7: iDAS Data Channel Properties  
PROPERTY  
NAME  
DESCRIPTION  
DEFAULT  
“NONE”  
SETTING RANGE  
The name of the data channel.  
Up to 6 letters and digits  
(more with APICOM, but  
only the first six are  
displayed on the front  
panel).  
TRIGGERING  
EVENT  
The event that triggers the data channel to measure  
and store its data parameters. See APPENDIX A-5  
for a list of available triggering events.  
ATIMER  
See Appendix A-5 For a  
complete list.  
NUMBER &  
PARAMETER  
LIST  
A User-configurable list of data types to be recorded  
in any given channel. See APPENDIX A-5 for a list  
of available parameters  
1 – DETMES  
See Appendix A-5 For a  
complete list.  
STARTING  
DATE  
The starting date when a channel starts collecting  
data  
01-JAN-03  
000:01:00  
Any actual date in the past  
or future.  
SAMPLE  
PERIOD  
The amount of time between each data point that is  
averaged into one mean reported every REPORT  
PERIOD.  
000:00:01 to 366:23:59  
(Days:Hours:Minutes)  
REPORT  
PERIOD  
The amount of time between each channel data  
point.  
000:01:00  
100  
000:00:01 to  
366:23:59  
(Days:Hours:Minutes)  
NUMBER OF  
RECORDS  
The number of reports that will be stored in the data  
file. Once the specified limit has been exceeded,  
the oldest data are over-written to make space for  
new data.  
1 to 1 million, limited by  
available storage space.  
RS-232  
REPORT  
Enables the analyzer to automatically report channel  
values to the RS-232 ports.  
OFF  
ON  
OFF or ON  
OFF or ON  
CHANNEL  
ENABLED  
Enables or disables the channel. Provides a  
convenient means to temporarily disable a data  
channel.  
CAL HOLD OFF Disables sampling of data parameters while  
OFF  
OFF or ON  
instrument is in calibration mode.  
(Section 6.7.2.11.)  
When enabled here – there is also a length of the  
DAS HOLD OFF after calibration mode, which is set  
in the VARS menu.  
60  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Model 360E Instruction Manual  
Operating Instructions  
6.7.1.2. iDAS Parameters  
Data parameters are types of data that may be measured and stored by the iDAS. For each Teledyne  
Instruments analyzer model, the list of available data parameters is different, fully defined and not customizable.  
Appendix A-5 lists firmware specific data parameters for the M360E. iDAS parameters include things like CO2  
concentration measurements, temperatures of the various heaters placed around the analyzer, pressures and  
flows of the pneumatic subsystem and other diagnostic measurements as well as calibration data such as slope  
and offset.  
Most data parameters have associated measurement units, such as mV, ppb, cm³/min, etc., although some  
parameters have no units. With the exception of concentration readings, none of these units of measure can be  
changed. To change the units of measure for concentration readings See Section 6.8.6.  
Note  
iDAS does not keep track of the unit of each concentration value and iDAS data files  
may contain concentrations in multiple units if the unit was changed during data  
acquisition.  
Each data parameter has user-configurable functions that define how the data are recorded:  
Table 6-8: iDAS Data Parameter Functions  
FUNCTION  
PARAMETER  
SAMPLE MODE  
EFFECT  
Instrument-specific parameter name.  
INST: Records instantaneous reading.  
AVG: Records average reading during reporting interval.  
MIN: Records minimum (instantaneous) reading during reporting interval.  
MAX: Records maximum (instantaneous) reading during reporting interval.  
SDEV: Records the standard deviation of the data points recorded during the reporting interval.  
Decimal precision of parameter value (0-4).  
PRECISION  
STORE NUM.  
SAMPLES  
OFF: stores only the average (default).  
ON: stores the average and the number of samples in each average for a parameter. This  
property is only useful when the AVG sample mode is used. Note that the number of samples  
is the same for all parameters in one channel and needs to be specified only for one of the  
parameters in that channel.  
Users can specify up to 50 parameters per data channel (the M360E provides about 30 parameters). However,  
the number of parameters and channels is ultimately limited by available memory.  
05232 Rev B3  
61  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Operating Instructions  
Model 360E Instruction Manual  
6.7.1.3. iDAS Triggering Events  
Triggering events define when and how the iDAS records a measurement of any given data channel. Triggering  
events are firmware-specific and a complete list of Triggers for this model analyzer can be found in Appendix A-  
5. The most commonly used triggering events are:  
ATIMER: Sampling at regular intervals specified by an automatic timer. Most trending  
information is usually stored at such regular intervals, which can be instantaneous or averaged.  
EXITZR, EXITSP, SLPCHG (exit zero, exit span, slope change): Sampling at the end of  
(irregularly occurring) calibrations or when the response slope changes. These triggering  
events create instantaneous data points, e.g., for the new slope and offset (concentration  
response) values at the end of a calibration. Zero and slope values are valuable to monitor  
response drift and to document when the instrument was calibrated.  
WARNINGS: Some data may be useful when stored if one of several warning messages  
appears such as WTEMPW (GFC wheel temperature warning) or PPRESW (purge pressure  
warning). This is helpful for trouble-shooting by monitoring when a particular warning occurred.  
6.7.2. Default iDAS Channels  
A set of default Data Channels has been included in the analyzer’s software for logging CO2 concentration and  
certain predictive diagnostic data. These default channels include but are not limited to:  
CONC: Samples CO2 concentration at one minute intervals and stores an average every hour with a time and  
date stamp. Readings during calibration and calibration hold off are not included in the data. By default, the last  
800 hourly averages are stored.  
PNUMTC: Collects sample flow and sample pressure data at five minute intervals and stores an average once a  
day with a time and date stamp. This data is useful for monitoring the condition of the pump and critical flow  
orifice (sample flow) and the sample filter (clogging indicated by a drop in sample pressure) over time to predict  
when maintenance will be required. The last 360 daily averages (about 1 year) are stored.  
CALDAT: Logs new slope and offset every time a zero or span calibration is performed. This Data Channel also  
records the instrument readings just prior to performing a calibration. This information is useful for performing  
predictive diagnostics as part of a regular maintenance schedule (See Section 9.1).  
STBZRO: Logs the concentration stability, the electronic output of the IR detector of the most recent measure  
phase and the measure/reference ratio every time the instrument exits the zero calibration mode. Data from the  
last 200 zero calibrations is stored. A time and date stamp is recorded for every data point logged. This  
information is useful for performing predictive diagnostics as part of a regular maintenance schedule (See  
Section 9.1).  
STBSPN: Logs the electronic output of the IR detector of the most recent measure phase and the  
measure/reference ratio every time the instrument exits span calibration mode. Data from the last 200 zero  
calibrations is stored. A time and date stamp is recorded for every data point logged. This information is useful  
for performing predictive diagnostics as part of a regular maintenance schedule (See Section 9.1).  
TEMP: Samples the analyzer’s bench temperature, box temperature and PHT cooler drive voltage every five  
minutes and records an average once every six hours. Data from the last 400 averaging periods is recorded. A  
time and date stamp is recorded for every data point logged. This information is useful for performing predictive  
diagnostics as part of a regular maintenance schedule (See Section 9.1).  
62  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Model 360E Instruction Manual  
Operating Instructions  
Note  
The CALDAT, STBZRO and STBSPN channels collect data based on events (e.g. a  
calibration operation) rather than a timed interval. This does not represent any specific  
length of time since it is dependent on how often calibrations are performed.  
Triggering Events and Data Parameters/Functions for these default channels are:  
LIST OF CHANNELS  
LIST OF PARAMETERS  
LIST OF PARAMETERS  
NAME: CONC  
EVENT: ATIMER  
PARAMETER: CONC1  
MODE: AVG  
PRECISION: 1  
REPORT PERIOD: 000:01:00  
NO. OF RECORDS: 800  
RS-232 REPORT: OFF  
CHANNEL ENABLED: ON  
CAL HOLD OFF: ON  
STORE NUM SAMPLES OFF  
PARAMETER: SMPLFLW  
MODE: AVG  
NAME: PNUMTC  
EVENT: ATIMER  
PRECISION: 1  
STORE NUM SAMPLES OFF  
REPORT PERIOD: 001:00:00  
NO. OF RECORDS: 360  
RS-232 REPORT: OFF  
CHANNEL ENABLED: ON  
CAL HOLD OFF: OFF  
PARAMETER: SMPLPRS  
MODE: AVG  
PARAMETER: SLOPE1  
MODE: INST  
PRECISION:3  
PRECISION: 1  
STORE NUM SAMPLES OFF  
STORE NUM SAMPLES OFF  
NAME: CALDAT  
EVENT: SLPCHG  
PARAMETER: OFSET1  
MODE: INST  
PRECISION: 1  
REPORT PERIOD: N/A  
NO. OF RECORDS:200  
RS-232 REPORT: OFF  
CHANNEL ENABLED: ON  
CAL HOLD OFF: OFF  
STORE NUM SAMPLES OFF  
PARAMETER: STABIL  
MODE: INST  
PARAMETER: ZSCNC1  
MODE: INST  
PRECISION:2  
STORE NUM SAMPLES OFF  
PRECISION: 1  
STORE NUM SAMPLES OFF  
NAME: STBZRO  
EVENT: EXITZR  
PARAMETER: DETMES  
MODE: INST  
PRECISION: 1  
STORE NUM SAMPLES OFF  
REPORT PERIOD: N/A  
NO. OF RECORDS:200  
RS-232 REPORT: OFF  
CHANNEL ENABLED: ON  
CAL HOLD OFF: OFF  
PARAMETER: DETMES  
MODE: INST  
PRECISION: 1  
PARAMETER: RATIO  
MODE: INST  
PRECISION: 3  
STORE NUM SAMPLES OFF  
NAME: STBSPN  
EVENT: EXITSP  
STORE NUM SAMPLES OFF  
REPORT PERIOD: N/A  
NO. OF RECORDS:200  
RS-232 REPORT: OFF  
CHANNEL ENABLED: ON  
CAL HOLD OFF: OFF  
PARAMETER: RATIO  
MODE: INST  
PRECISION: 3  
STORE NUM SAMPLES OFF  
PARAMETER: BNTEMP  
MODE: AVG  
PRECISION:1  
STORE NUM SAMPLES OFF  
NAME: TEMP  
EVENT: ATIMER  
PARAMETER: BOXTMP  
MODE: AVG  
PRECISION: 1  
STORE NUM SAMPLES OFF  
REPORT PERIOD: 000:06:00  
NO. OF RECORDS:400  
RS-232 REPORT: OFF  
CHANNEL ENABLED: ON  
CAL HOLD OFF: OFF  
PARAMETER: PHTDRV  
MODE: AVG  
PRECISION: 1  
STORE NUM SAMPLES OFF  
Figure 6-4: Default iDAS Channels Setup  
These default Data Channels can be used as they are, or they can be customized from the front panel to fit a  
specific application. They can also be deleted to make room for custom user-programmed Data Channels.  
Appendix A-5 lists the firmware-specific iDAS configuration in plain-text format. This text file can either be  
loaded into APICOM and then modified and uploaded to the instrument or can be copied and pasted into a  
terminal program to be sent to the analyzer.  
05232 Rev B3  
63  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Operating Instructions  
Model 360E Instruction Manual  
NOTE  
Sending an iDAS configuration to the analyzer through its COM ports will replace the  
existing configuration and will delete all stored data. Back up any existing data and the  
iDAS configuration before uploading new settings.  
These default Data Channels can be used as they are, or they can be customized from the front panel to fit a  
specific application. They can also be deleted to make room for custom user-programmed Data Channels.  
Appendix A-5 lists the firmware-specific iDAS configuration in plain-text format. This text file can either be  
loaded into APICOM and then modified and uploaded to the instrument or can be copied and pasted into a  
terminal program to be sent to the analyzer.  
NOTE  
Sending an iDAS configuration to the analyzer through its COM ports will replace the  
existing configuration and will delete all stored data. Back up any existing data and the  
iDAS configuration before uploading new settings.  
64  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual  
Operating Instructions  
6.7.2.1. Viewing iDAS Data and Settings  
iDAS data and settings can be viewed on the front panel through the following keystroke sequence.  
VIEW KEYPAD FUNCTIONS  
KEY  
FUNCTION  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
<PRM  
PRM>  
Moves to the next Parameter  
Moves to the previous  
Parameter  
SETUP X.X  
PRIMARY SETUP MENU  
NX10  
NEXT  
PREV  
PV10  
Moves the view forward 10  
data points/channels  
CFG DAS RNGE PASS CLK MORE  
EXIT  
EXIT  
Moves to the next data  
point/channel  
Moves to the previous data  
point/channel  
SETUP X.X  
DATA ACQUISITION  
Moves the view back 10 data  
points/channels  
VIEW EDIT  
Keys only appear as needed  
SETUP X.X  
CONC : DATA AVAILABLE  
NEXT VIEW  
EXIT  
SETUP X.X  
00:00:00 NXCNC1=0.0 PPM  
PV10 PREV NEXT NX10 <PRM PRM>  
EXIT  
SETUP X.X  
PNUMTC: DATA AVAILABLE  
PREV NEXT VIEW  
EXIT  
SETUP X.X  
00:00:00 SMPFLW=000.0 cc / m  
<PRM  
PRM>  
EXIT  
SETUP X.X  
CALDAT: DATA AVAILABLE  
PREV NEXT VIEW  
EXIT  
SETUP X.X  
00:00:00 NXSLP1=0.000  
<PRM PRM>  
PV10 PREV  
EXIT  
SETUP X.X  
STBZRO: DATA AVAILABLE  
PREV NEXT VIEW  
EXIT  
SETUP X.X  
00:00:00 STABIL=0.000  
<PRM PRM>  
PV10 PREV  
EXIT  
EXIT  
EXIT  
SETUP X.X  
STBSPN: DATA AVAILABLE  
PREV NEXT VIEW  
EXIT  
SETUP X.X  
00:00:00 DETMES=0.000  
<PRM PRM>  
PV10 PREV  
SETUP X.X  
TEMP: DATA AVAILABLE  
PREV NEXT VIEW  
EXIT  
SETUP X.X  
00:00:00 BOXTMP=0.000  
<PRM PRM>  
PV10 PREV  
05232 Rev B3  
65  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Operating Instructions  
Model 360E Instruction Manual  
6.7.2.2. Editing iDAS Data Channels  
iDAS configuration is most conveniently done through the APICOM remote control program. The following list of  
key strokes shows how to edit using the front panel.  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
SETUP X.X  
PRIMARY SETUP MENU  
EXIT will return to the  
previous SAMPLE  
display.  
CFG DAS RNGE PASS CLK MORE  
EXIT  
Main Data Acquisition Menu  
SETUP X.X  
DATA ACQUISITION  
VIEW EDIT  
EXIT  
SAMPLE  
ENTER SETUP PASS : 818  
8
8
1
ENTR EXIT  
Edit Data Channel Menu  
Moves the  
display up &  
down the list of  
Data Channels  
SETUP X.X  
0) CONC: ATIMER, 1,  
900  
Exits to the Main  
Data Acquisition  
Menu  
PREV NEXT  
INS DEL EDIT PRNT EXIT  
Exports the  
Inserts a new Data  
Channel into the list  
BEFORE the Channel  
currently being displayed  
configuration of all  
data channels to  
RS-232 interface.  
Deletes The Data  
Channel currently  
being displayed  
Moves the display  
between the  
SETUP X.X  
NAME:CONC  
Exits returns to the  
previous Menu  
PROPERTIES for this  
data channel.  
<SET SET> EDIT PRNT  
EXIT  
Reports the configuration of current  
data channels to the RS-232 ports.  
Allows to edit the channel name, see next key sequence.  
When editing the data channels, the top line of the display indicates some of the configuration parameters. For  
example, the display line:  
0) CONC: ATIMER, 4, 800  
translates to the following configuration:  
Channel No.: 0  
NAME: CONC  
TRIGGER EVENT: ATIMER  
PARAMETERS: Four parameters are included in this channel  
EVENT: This channel is set up to record 800 data points.  
To edit the name of a data channel, follow the above key sequence and then press:  
66  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 360E Instruction Manual  
Operating Instructions  
From the end of the previous key sequence …  
SETUP X.X  
NAME:CONC  
<SET SET> EDIT PRINT  
EXIT  
ENTR accepts the new string  
SETUP X.X  
NAME:CONC  
and returns to the previous  
menu.  
C
O
N
C
-
-
ENTR  
EXIT  
EXIT ignores the new string  
and returns to the previous  
menu.  
Press each key repeatedly to cycle through the  
available character set:  
0-9, A-Z, space ’ ~ ! # $ % ^ & * ( ) - _ = +[ ]  
{ } < >\ | ; : , . / ?  
6.7.2.3. Trigger Events  
To edit the list of data parameters associated with a specific data channel, press:  
From the DATA ACQUISITION menu  
(see Section 6.7.2.2)  
Edit Data Channel Menu  
SETUP X.X  
0) CONC: ATIMER, 1,  
900  
Exits to the Main  
Data Acquisition  
menu  
PREV NEXT  
INS DEL EDIT PRNT EXIT  
SETUP X.X  
NAME:CONC  
<SET SET> EDIT PRINT  
EXIT  
SETUP X.X  
EVENT:ATIMER  
<SET SET> EDIT PRINT  
EXIT  
ENTR accepts the new string  
and returns to the previous  
menu.  
EXIT ignores the new string  
and returns to the previous  
menu.  
SETUP X.X  
EVENT:ATIMER  
<PREV NEXT>  
ENTR  
EXIT  
Press each key repeatedly to cycle through the  
list of available trigger events.  
05232 Rev B3  
67  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Operating Instructions  
Model 360E Instruction Manual  
6.7.2.4. Editing iDAS Parameters  
Data channels can be edited individually from the front panel without affecting other data channels. However,  
when editing a data channel, such as during adding, deleting or editing parameters, all data for that particular  
channel will be lost, because the iDAS can store only data of one format (number of parameter columns etc.) for  
any given channel. In addition, an iDAS configuration can only be uploaded remotely as an entire set of  
channels. Hence, remote update of the iDAS will always delete all current channels and stored data.  
To modify, add or delete a parameter, follow the instruction shown in Section 6.7.2.2 then press:  
From the DATA ACQUISITION menu  
(see Section 6.7.2.2)  
Edit Data Channel Menu  
SETUP X.X  
0) CONC: ATIMER, 1,  
900  
Exits to the main  
Data Acquisition  
menu  
PREV NEXT  
INS DEL EDIT PRNT EXIT  
SETUP X.X  
NAME:CONC  
<SET SET> EDIT PRINT  
EXIT  
Press SET> key until…  
SETUP X.X  
PARAMETERS:1  
<SET SET> EDIT PRINT  
EXIT  
SETUP X.X  
EDIT PARAMS (DELETE DATA)  
YES will delete  
all data in that  
entire channel.  
NO returns to  
the previous  
menu and  
YES NO  
retains all data.  
Edit Data Parameter Menu  
Moves the  
display between  
existing  
SETUP X.X 0) PARAM=CONC1, MODE=AVG  
PREV NEXT INS DEL EDIT  
Exits to the main  
Data Acquisition  
menu  
EXIT  
Parameters  
Inserts a new Parameter  
before the currently  
displayed Parameter  
Use to configure  
the functions for  
this Parameter.  
Deletes the Parameter  
currently displayed.  
68  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 360E Instruction Manual  
Operating Instructions  
To configure a specific data parameter, press:  
FROM THE EDIT DATA PARAMETER MENU  
(see previous section)  
SETUP X.X 0) PARAM=CONC1, MODE=AVG  
PREV NEXT  
INS DEL EDIT  
EXIT  
SETUP X.X PARAMETERS:CONC!  
SET> EDIT  
EXIT  
SETUP X.X PARAMETERS: PMTDET  
PREV NEXT ENTR  
EXIT  
If more than on parameter is active for  
this channel, these cycle through list of  
existing Parameters.  
SETUP X.X SAMPLE MODE:AVG  
<SET SET> EDIT  
EXIT  
SETUP X.X SAMPLE MODE: INST  
INST AVG MIN MAX  
EXIT  
Press the key for the desired mode  
ENTR accepts the new  
setting and returns to the  
previous menu.  
SETUP X.X PRECISION: 1  
EXIT ignores the new setting  
and returns to the previous  
<SET SET> EDIT  
EXIT  
SETUP X.X PRECISION: 1  
1
EXIT  
Set for 0-4  
<SET Returns to  
previous  
SETUP X.X STORE NUM. SAMPLES: OFF  
Functions  
<SET  
EDIT  
EXIT  
SETUP X.X STORE NUM. SAMPLES: OFF  
OFF  
ENTR EXIT  
Turn ON or OFF  
05232 Rev B3  
69  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operating Instructions  
Model 360E Instruction Manual  
6.7.2.5. Sample Period and Report Period  
The iDAS defines two principal time periods by which sample readings are taken and permanently recorded:  
SAMPLE PERIOD: Determines how often iDAS temporarily records a sample reading of the parameter  
in volatile memory. The SAMPLE PERIOD is set to one minute by default and generally cannot be  
accessed from the standard iDAS front panel menu, but is available via the instruments communication  
ports by using APICOM or the analyzer’s standard serial data protocol.  
SAMPLE PERIOD is only used when the iDAS parameter’s sample mode is set for AVG, MIN or MAX.  
REPORT PERIOD: Sets how often the sample readings stored in volatile memory are processed, (e.g.  
average, minimum or maximum are calculated) and the results stored permanently in the instruments  
Disk-on-Chip as well as transmitted via the analyzer’s communication ports. The REPORT PERIOD  
may be set from the front panel.  
If the INST sample mode is selected the instrument stores and reports an instantaneous reading of the  
selected parameter at the end of the chosen REPORT PERIOD  
In AVG, MIN or MAX sample modes, the settings for the SAMPLE PERIOD and the REPORT PERIOD  
determine the number of data points used each time the average, minimum or maximum is calculated, stored  
and reported to the COMM ports. The actual sample readings are not stored past the end of the of the chosen  
REPORT PERIOD.  
Also, the SAMPLE PERIOD and REPORT PERIOD intervals are synchronized to the beginning and end of the  
appropriate interval of the instruments internal clock.  
If SAMPLE PERIOD were set for one minute the first reading would occur at the beginning of the next  
full minute according to the instrument’s internal clock.  
If the REPORT PERIOD were set for of one hour the first report activity would occur at the beginning of  
the next full hour according to the instrument’s internal clock.  
EXAMPLE: Given the above settings, if iDAS were activated at 7:57:35 the first sample would occur at  
7:58 and the first report would be calculated at 8:00 consisting of data points for 7:58. 7:59 and 8:00.  
During the next hour (from 8:01 to 9:00) the instrument will take a sample reading every minute and  
include 60 sample readings.  
When the STORE NUM. SAMPLES feature is turned on the instrument will also store how many sample  
readings were used for the AVG, MIN or MAX calculation but not the readings themselves.  
REPORT PERIODS IN PROGRESS WHEN INSTRUMENT IS POWERED OFF  
If the instrument is powered off in the middle of a REPORT PERIOD, the samples accumulated so far during that  
period are lost. Once the instrument is turned back on, the iDAS restarts taking samples and temporarily them  
in volatile memory as part of the REPORT PERIOD currently active at the time of restart. At the end of this  
REPORT PERIOD only the sample readings taken since the instrument was turned back on will be included in  
any AVG, MIN or MAX calculation. Also, the STORE NUM. SAMPLES feature will report the number of sample  
readings taken since the instrument was restarted.  
70  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 360E Instruction Manual  
Operating Instructions  
To define the REPORT PERIOD, follow the instruction shown in Section 6.7.2.2 then press:  
From the DATA ACQUISITION menu  
(see Section 6.7.2.2)  
Edit Data Channel Menu  
Use the PREV and NEXT  
keys to scroll to the data  
channel to be edited.  
SETUP X.X  
0) CONC: ATIMER, 1,  
900  
Exits to the main  
Data Acquisition  
menu.  
PREV NEXT  
INS DEL EDIT PRNT EXIT  
SETUP X.X  
NAME:CONC  
<SET SET> EDIT PRINT  
EXIT  
Press SET> key until you reach REPORT PERIOD …  
SETUP X.X  
REPORT PERIOD:000:01:00  
<SET SET> EDIT PRINT  
EXIT  
SETUP X.X  
REPORT PERIODD:DAYS:0  
Set the number of days  
between reports (0-366).  
0
0
0
ENTR EXIT  
Press keys to set hours  
between reports in the format :  
HH:MM (max: 23:59). This is a  
24 hour clock . PM hours are 13  
thru 23, midnight is 00:00.  
SETUP X.X  
REPORT PERIODD:TIME:01:01  
ENTR EXIT  
ENTR accepts the new string and  
returns to the previous menu.  
EXIT ignores the new string and  
returns to the previous menu.  
0
1
0
0
IIf at any time an illegal entry is selected (e.g., days > 366)  
the ENTR key will disappear from the display.  
Example 2:15 PM = 14:15  
05232 Rev B3  
71  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operating Instructions  
Model 360E Instruction Manual  
6.7.2.6. Number of Records  
The number of data records in the M360E is limited to about a cumulative one million data points in all channels  
(one megabyte of space on the disk-on-chip). However, the actual number of records is also limited by the total  
number of parameters and channels and other settings in the iDAS configuration. Every additional data channel,  
parameter, number of samples setting etc. and will reduce the maximum amount of data points somewhat. In  
general, however, the maximum data capacity is divided amongst all channels (max: 20) and parameters (max:  
50 per channel).  
The iDAS will check the amount of available data space and prevent the user from specifying too many records  
at any given point. If, for example, the iDAS memory space can accommodate 375 more data records, the  
ENTR key will disappear when trying to specify more than that number of records. This check for memory space  
may also make an upload of an iDAS configuration with APICOM or a Terminal program fail, if the combined  
number of records would be exceeded. In this case, it is suggested to either try from the front panel what the  
maximum number of records can be or use trial-and-error in designing the iDAS script or calculate the number of  
records using the DAS or APICOM manuals. To set the number of records for one channel from the front panel,  
press SETUP-DAS-EDIT-ENTR and the following key sequence.  
From the DATA ACQUISITION menu  
(see Section 6.12.2.2)  
Edit Data Channel Menu  
SETUP X.X  
0) CONC: ATIMER, 1 2,  
900  
Exits to the main  
Data Acquisition  
menu  
PREV NEXT  
INS DEL EDIT PRNT EXIT  
SETUP X.X  
NAME:CONC  
<SET SET> EDIT PRINT  
EXIT  
Press SET> key until…  
SETUP X.X  
NUMBER OF RECORDS:000  
<SET SET> EDIT PRINT  
EXIT  
SETUP X.X  
EDIT RECOPRDS (DELET DATA)  
NO returns to the  
previous menu.  
YES will delete all data  
in this channel.  
YES  
NO  
ENTR accepts the new  
setting and returns to the  
previous menu.  
EXIT ignores the new setting  
and returns to the previous  
menu.  
Toggle keys to set  
number of records  
(1-99999)  
SETUP X.X  
REPORT PERIODD:DAYS:0  
ENTR EXIT  
0
0
0
0
0
72  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 360E Instruction Manual  
Operating Instructions  
6.7.2.7. RS-232 Report Function  
The M360E iDAS can automatically report data to the communications ports, where they can be captured with a  
terminal emulation program or simply viewed by the user.  
To enable automatic COM port reporting, follow the instruction shown in Section 6.7.2.2 then press:  
From the DATA ACQUISITION menu  
(see Section 6.7.2.2)  
Edit Data Channel Menu  
SETUP X.X  
0) CONC: ATIMER, 1,  
900  
Exits to the main  
Data Acquisition  
menu  
PREV NEXT  
INS DEL EDIT PRNT EXIT  
SETUP X.X  
NAME:CONC  
<SET SET> EDIT PRINT  
EXIT  
Press SET> key until…  
SETUP X.X  
RS-232 REPORT: OFF  
<SET SET> EDIT PRINT  
EXIT  
ENTR accepts the new  
setting and returns to the  
previous menu.  
EXIT ignores the new setting  
and returns to the previous  
menu.  
SETUP X.X  
RS-232 REPORT: OFF  
Toggle key to turn  
reporting ON or OFF  
OFF  
ENTR EXIT  
6.7.2.8. Compact Report  
When enabled, this option avoids unnecessary line breaks on all RS-232 reports. Instead of reporting each  
parameter in one channel on a separate line, up to five parameters are reported in one line.  
6.7.2.9. Starting Date  
This option allows to specify a starting date for any given channel in case the user wants to start data acquisition  
only after a certain time and date. If the Starting Date is in the past, the iDAS ignores this setting.  
05232 Rev B3  
73  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Operating Instructions  
Model 360E Instruction Manual  
6.7.2.10. Disabling/Enabling Data Channels  
Data channels can be temporarily disabled, which can reduce the read/write wear on the disk-on-chip. The  
ALL_01 channel of the M360E, for example, is disabled by default.  
To disable a data channel, follow the instruction shown in Section 6.7.2.2 then press:  
From the DATA ACQUISITION menu  
(see Section 6.7.2.2)  
Edit Data Channel Menu  
SETUP X.X  
0) CONC: ATIMER, 1,  
900  
Exits to the main  
Data Acquisition  
menu  
PREV NEXT  
INS DEL EDIT PRNT EXIT  
SETUP X.X  
NAME:CONC  
<SET SET> EDIT PRINT  
EXIT  
Press SET> key until…  
SETUP X.X  
CHANNEL ENABLE:ON  
<SET SET> EDIT PRINT  
EXIT  
ENTR accepts the new  
setting and returns to the  
previous menu.  
EXIT ignores the new setting  
and returns to the previous  
menu.  
SETUP X.X  
CHANNEL ENABLE:ON  
Toggle key to turn  
channel ON or OFF  
OFF  
ENTR EXIT  
6.7.2.11. HOLDOFF Feature  
The iDAS HOLDOFF feature allows to prevent data collection during calibrations and during the  
DAS_HOLDOFF period enabled and specified in the VARS (Section 6.12). To enable or disable the HOLDOFF,  
follow the instruction shown in Section 6.7.2.2 then press:  
From the DATA ACQUISITION menu  
(see Section 6.7.2.2)  
Edit Data Channel Menu  
SETUP X.X  
0) CONC: ATIMER, 1,  
900  
Exits to the main  
Data Acquisition  
menu  
PREV NEXT  
INS DEL EDIT PRNT EXIT  
SETUP X.X  
NAME:CONC  
<SET SET> EDIT PRINT  
EXIT  
Press SET> key until…  
SETUP X.X  
CAL HOLD OFF:ON  
SET> EDIT PRINT  
EXIT  
ENTR accepts the new  
setting and returns to the  
previous menu.  
EXIT ignores the new setting  
and returns to the previous  
menu.  
SETUP X.X  
CAL HOLD OFF:ON  
Toggle key to turn  
HOLDOFF ON or OFF  
ON  
ENTR EXIT  
74  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Model 360E Instruction Manual  
Operating Instructions  
6.7.3. Remote iDAS Configuration  
Editing channels, parameters and triggering events as described in this can be performed via the APICOM  
remote control program using the graphic interface shown in Figure 6-5. Refer to Section 6.15 for details on  
remote access to the M360E analyzer.  
Figure 6-5: APICOM user interface for configuring the iDAS.  
Once an iDAS configuration is edited (which can be done offline and without interrupting DAS data collection), it  
is conveniently uploaded to the instrument and can be stored on a computer for later review, alteration or  
documentation and archival. Refer to the APICOM manual for details on these procedures. The APICOM user  
manual (Teledyne Instruments part number 039450000) is included in the APICOM installation file, which can be  
downloaded at http://www.teledyne-api.com/software/apicom/.  
Although Teledyne Instruments recommends the use of APICOM, the iDAS can also be accessed and  
configured through a terminal emulation program such as HyperTerminal (Figure 6-6). However, all  
configuration commands must be created following a strict syntax or be pasted in from of a text file, which was  
edited offline and then uploaded through a specific transfer procedure.  
05232 Rev B3  
75  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Operating Instructions  
Model 360E Instruction Manual  
Figure 6-6: iDAS Configuration Through a Terminal Emulation Program.  
Both procedures are best started by downloading the default iDAS configuration, getting familiar with its  
command structure and syntax conventions, and then altering a copy of the original file offline before uploading  
the new configuration.  
CAUTION  
Whereas the editing, adding and deleting of iDAS channels and parameters of one  
channel through the front-panel keyboard can be done without affecting the other  
channels, uploading an iDAS configuration script to the analyzer through its  
communication ports will erase all data, parameters and channels by replacing them  
with the new iDAS configuration. Backup of data and the original iDAS configuration is  
advised before attempting any iDAS changes.  
76  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 360E Instruction Manual  
Operating Instructions  
6.8. SETUP RNGE: Analog Output Reporting Range  
Configuration  
The analyzer has three active analog output signals, accessible through a connector on the rear panel.  
ANALOG OUT  
CO2 concentration  
Test Channel  
outputs  
Not Used  
A1  
A2  
A3  
A4  
+
-
+
-
+
-
+
-
LOW range when DUAL  
HIGH range when DUAL  
mode is selected  
mode is selected  
Figure 6-7: Analog Output Connector Pin Out  
All three outputs can be configured either at the factory or by the user for full scale outputs of 0.1 VDC, 1VDC,  
5VDC or 10VDC. Additionally A1 and A2 may be equipped with optional 0-20 mADC current loop drivers and  
configured for any current output within that range (e.g. 0-20, 2-20, 4-20, etc.). The user may also adjust the  
signal level and scaling of the actual output voltage or current to match the input requirements of the recorder or  
data logger (See Section 6.13.4).  
The A1 and A2 channels output a signal that is proportional to the CO2 concentration of the sample gas.  
Several modes are available which allow them to operate independently or be slaved together (See Section 6.7).  
The user may also select between a variety of reporting range spans (See Sections 6.8.3, 6.8.4 and 6.8.5).  
EXAMPLE:  
A1 OUTPUT: Output Signal = 0-5 VDC representing 0-1000 ppm concentration values  
A2 OUTPUT: Output Signal = 0 – 10 VDC representing 0-500 ppm concentration values.  
Output A3 is only active if the O2 sensor option is installed. In this case a signal representing the currently  
measured O2 concentration is output on this channel.  
The output, labeled A4 is special. It can be set by the user (See Section 6.13.9) to output several of the test  
functions accessible through the <TST TST> keys of the units sample display.  
6.8.1. Physical Range versus Analog Output Reporting Ranges  
Functionally, the Model 360E Gas Filter Correlation CO2 Analyzer has one hardware Physical Range that is  
capable of determining CO2 concentrations between 50 ppb and 2,000 ppm. This architecture improves  
reliability and accuracy by avoiding the need for extra, switchable, gain-amplification circuitry. Once properly  
calibrated, the analyzer’s front panel will accurately report concentrations along the entire span of its 50 ppb and  
2,000 ppm physical range.  
Because, most applications use only a small part of the analyzer’s physical range, the width of the Model 360E’s  
physical range can create data resolution problems for most analog recording devices. For example, in an  
application where the expected concentration of CO2 is typically less than 500 ppm, the full scale of expected  
values is only 25% of the instrument’s 2,000 ppm physical range. Unmodified, the corresponding output signal  
would also be recorded across only 25% of the range of the recording device.  
05232 Rev B3  
77  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Operating Instructions  
Model 360E Instruction Manual  
The M360E solves this problem by allowing the user to select a scaled reporting range for the analog outputs  
that only includes that portion of the physical range relevant to the specific application. Only the reporting range  
of the analog outputs is scaled, the physical range of the analyzer and the readings displayed on the front panel  
remain unaltered.  
6.8.2. Reporting Range Modes  
The M360E provides three analog output range modes to choose from.  
Single range (SNGL) mode sets a single maximum range for the analog output. If single range is  
selected (See Section 6.78.3) both outputs are slaved together and will represent the same  
measurement span (e.g. 0-50 ppm), however their electronic signal levels may be configured for  
different ranges (e.g. 0-10 VDC vs. 0-.1 VDC – See Section 6.9.4.1).  
Dual range (DUAL) allows the A1 and A2 outputs to be configured with different measurement spans  
(See Section 6.8.4) as well as separate electronic signal levels (See Section 6.9.4.1).  
Auto range (AUTO) mode gives the analyzer to ability to output data via a low range and high range.  
When this mode is selected (See Section 6.8.5) the M360E will automatically switch between the two  
ranges dynamically as the concentration value fluctuates.  
Range status is also output via the external digital I/O status outputs (See Section 6.15.1.1).  
To select the Analog Output Range Type press:  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
SAMPLE  
ENTER SETUP PASS : 818  
8
1
8
ENTR EXIT  
SETUP X.X  
CFG DAS RNGE PASS CLK MORE  
EXIT  
SETUP X.X  
RANGE CONTROL MENU  
MODE SET UNIT  
EXIT  
SETUP X.X  
RANGE MODE: SNGL  
EXIT Returns  
to the Main  
SNGL DUAL AUTO  
ENTR EXIT  
SAMPLE Display  
Only one of the  
range modes may  
be active at any  
time.  
Go To  
Section  
6.7.3  
Go To  
Section  
6.7.4  
Go To  
Section  
6.7.5  
78  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 360E Instruction Manual  
Operating Instructions  
NOTE  
Upper span limit setting for the individual range modes are shared. Resetting the span  
limit in one mode also resets the span limit for the corresponding range in the other  
modes as follows:  
SNGL  
Range  
DUAL  
AUTO  
 Range1  Low Range  
Range2  High Range  
6.8.3. Single Range mode (SNGL)  
This is the default reporting range mode for the analyzer. In single range mode both A1 and A2 are set to the  
same reporting range. This reporting range can be any value between 50 ppb and 2 000 ppm.  
While the two outputs always have the same reporting range, the span, signal offset and scaling of their  
electronic signals may be configured for differently (e.g., A1 = 0-10 V; A2 = 0-0.1 V). See Section 6.13.4 for  
instructions on adjusting these parameters.  
To select SNGL range mode and to set the upper limit of the range, press:  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 X.XXX  
SETUP C.3  
RANGE MODE: SNGL  
< TST TST > CAL  
SETUP  
SNGL DUAL AUTO  
ENTR EXIT  
SETUP C.3  
RANGE CONTROL MENU  
SETUP C.3  
PRIMARY SETUP MENU  
MODE SET UNIT  
EXIT  
CFG DAS RNGE PASS CLK MORE  
EXIT  
SETUP C.3  
RANGE CONTROL MENU  
SETUP C.3  
RANGE: 500.0 Conc  
.0  
MODE SET UNIT  
EXIT  
0
0
5
0
0
ENTR EXIT  
SETUP C.3  
RANGE MODE: SNGL  
SETUP C.3  
MODE SET UNIT  
RANGE CONTROL MENU  
EXIT x 2 returns  
to the main  
SAMPLE display  
SNGL DUAL AUTO  
ENTR EXIT  
EXIT  
05232 Rev B3  
79  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Operating Instructions  
Model 360E Instruction Manual  
6.8.4. Dual Range Mode (DUAL)  
Selecting Dual Range mode allows the A1 and A2 outputs to be configured with different reporting ranges. The  
analyzer software calls these two ranges low and high. The low range setting corresponds with the analog  
output labeled A1 on the Rear Panel of the instrument. The high Range Setting corresponds with the A2 output.  
While the software names these two ranges low and high, they do not have to be configured that way. For  
example: The low range can be set for a span of 0-1000 ppm while the high range is set for 0-500 ppm.  
In DUAL range mode the RANGE test function displayed on the front panel will be replaced by two separate  
functions:  
RANGE1: The range setting for the A1 output.  
RANGE2: The range setting for the A2 output.  
To set the ranges press following keystroke sequence  
SETUP X.X  
RANGE MODE: DUAL  
SNGL DUAL AUTO  
ENTR EXIT  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
SETUP X.X  
RANGE CONTROL MENU  
< TST TST > CAL  
SETUP  
MODE SET UNIT  
EXIT  
SETUP X.X  
PRIMARY SETUP MENU  
SETUP X.X  
LOW RANGE: 500.0 Conc  
.0 ENTR EXIT  
Toggle the  
Numeral Keys  
to set the upper  
limit of each  
range.  
CFG DAS RNGE PASS CLK MORE  
EXIT  
0
0
1
5
0
0
SETUP X.X  
RANGE CONTROL MENU  
SETUP X.X  
HIGH RANGE: 500.0 Conc  
.0 ENTR EXIT  
MODE SET UNIT  
EXIT  
0
0
0
0
SETUP X.X  
RANGE MODE: SNGL  
SETUP X.X  
RANGE CONTROL MENU  
EXIT Returns  
to the Main  
SAMPLE Display  
SNGL DUAL AUTO  
ENTR EXIT  
MODE SET UNIT  
EXIT  
.
When the instrument’s range mode is set to DUAL the concentration field in the upper right hand corner of the  
display alternates between displaying the low range value and the high range value. The concentration currently  
being displayed is identified as follows: C1 = Low (or A1) and C2 = High (or A2).  
NOTE  
In DUAL range mode the LOW and HIGH ranges have separate slopes and offsets for  
computing CO2 concentration.  
The two ranges must be independently calibrated.  
80  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 360E Instruction Manual  
Operating Instructions  
6.8.5. Auto Range Mode (AUTO)  
In AUTO range mode, the analyzer automatically switches the reporting range between two user-defined ranges  
(low and high). The unit will switch from low range to high range when the CO2 concentration exceeds 98% of  
the low range span. The unit will return from high range back to low range once both the CO2 concentration falls  
below 75% of the low range span.  
In AUTO Range mode the instrument reports the same data in the same range on both the A1 and A2 outputs  
and automatically switches both outputs between ranges as described above. Also, the RANGE test function  
displayed on the front panel will automatically switch to show which range is in effect.  
The high/low range status is also reported through the external, digital status outputs (Section 6.15.1.1).  
To set individual ranges press the following keystroke sequence.  
SETUP X.X  
RANGE MODE: AUTO  
SNGL DUAL AUTO  
ENTR EXIT  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
SETUP X.X  
RANGE CONTROL MENU  
EXIT x 2 returns  
to the main  
SAMPLE display  
MODE SET UNIT  
EXIT  
SETUP X.X  
PRIMARY SETUP MENU  
CFG DAS RNGE PASS CLK MORE  
EXIT  
EXIT  
SETUP X.X  
LOW RANGE: 500.0 Conc  
.0 ENTR EXIT  
Toggle the numeral  
keys to set the  
LOW and HIGH  
range value.  
SETUP X.X  
RANGE CONTROL MENU  
0
0
5
5
0
0
MODE SET UNIT  
ENTR accepts the  
new setting, EXIT  
ignores the new  
setting.  
SETUP X.X  
RANGE MODE: SNGL  
SETUP X.X  
HIGH RANGE: 500.0 Conc  
.0 ENTR EXIT  
SNGL DUAL AUTO  
ENTR EXIT  
0
0
0
0
CAUTION  
In AUTO range mode the LOW and HIGH ranges have separate slopes and offsets for  
computing CO2 concentration.  
The two ranges must be independently calibrated.  
NOTE  
Avoid accidentally setting the low range of the instrument with a higher span limit than  
the high range. This will cause the unit to stay in the low reporting range perpetually  
and defeat the function of the AUTO range mode.  
05232 Rev B3  
81  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Operating Instructions  
Model 360E Instruction Manual  
6.8.6. Range Units  
The M360E can display concentrations in parts per billion (109 mols per mol, PPB), parts per million (106 mols  
per mol, PPM), micrograms per cubic meter (µg/m3, UG), milligrams per cubic meter (mg/m3, MG) or percent  
(volume CO2/volume sample gas, %). Changing units affects all of the display, analog outputs, COM port and  
iDAS values for all reporting ranges regardless of the analyzer’s range mode.  
NOTE  
Concentrations displayed in mg/m3 and ug/m3 use 0C, 760 mmHg for Standard  
Temperature and Pressure (STP). Consult your local regulations for the STP used by  
your agency.  
Conversion factors from volumetric to mass units are:  
CO2: ppb x 1.96 = µg/m3; ppm x 1.96 = mg/m3  
To change the concentration units:  
SAMPLE  
RANGE = 500.00 PPB  
CO2=X.XXX  
< TST TST > CAL  
SETUP  
SETUP X.X  
PRIMARY SETUP MENU  
CFG DAS RNGE PASS CLK MORE  
EXIT  
EXIT  
EXIT returns  
to the main menu.  
SETUP X.X  
RANGE CONTROL MENU  
MODE SET UNIT  
SETUP X.X  
CONC UNITS: PPM  
Select the preferred  
concentration unit.  
PPM PPB UGM MGM  
%
ENTER EXIT  
ENTER EXIT  
ENTR accepts  
the new unit,  
EXIT returns  
to the SETUP  
menu.  
SETUP X.X  
CONC UNITS: %  
PPM PPB UGM MGM  
%
NOTE  
Once the units of measurement have been changed the unit MUST be recalibrated, as  
the “expected span values” previously in effect will no longer be valid. Simply entering  
new expected span values without running the entire calibration routine is not  
sufficient.  
82  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 360E Instruction Manual  
Operating Instructions  
6.8.7. Dilution Ratio  
The dilution ratio is a software option that allows the user to compensate for any dilution of the sample gas  
before it enters the sample inlet. Using the dilution ratio option is a 4-step process:  
1. Select reporting range units: Follow the procedure in Section 6.8.6.  
2. Select the range: Use the procedures in Sections 6.8.2 – 6.8.5. Make sure that the SPAN value entered  
is the maximum expected concentration of the undiluted calibration gas and that the span gas is either  
supplied through the same dilution inlet system as the sample gas or has an appropriately lower actual  
concentration. For example, with a dilution set to 100, a 10 ppm gas can be used to calibrate a 1000  
ppm sample gas if the span gas is not routed through the dilution system. On the other hand, if a 1000  
ppm span gas is used, it needs to pass through the same dilution steps as the sample gas.  
3. Set the dilution factor as a gain (e.g., a value of 20 means 20 parts diluting gas and 1 part of sample  
gas):  
SAMPLE  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
SETUP C.3  
PRIMARY SETUP MENU  
CFG DAS RNGE PASS CLK MORE  
EXIT  
EXIT  
SETUP C.3  
RANGE CONTROL MENU  
DIL only appears  
if the dilution ratio  
option has been  
installed  
MODE SET UNIT DIL  
EXIT ignores the  
new setting.  
SETUP C.3  
DIL FACTOR: 1.0 GAIN  
.0 ENTR  
ENTR accepts the  
Toggle these keys to set the dilution  
factor.  
new setting.  
0
0
0
1
EXIT  
This is the number by which the  
analyzer will multiply the CO2  
concentrations of the gas passing  
through the reaction cell.  
SETUP C.3  
DIL FACTOR: 20.0 GAIN  
.0 ENTR  
0
0
2
0
EXIT  
The analyzer multiplies the measured gas concentrations with this dilution factor and displays the result.  
NOTE  
Once the above settings have been entered, the instrument needs to be recalibrated  
using one of the methods discussed in Chapter 7.  
05232 Rev B3  
83  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Operating Instructions  
Model 360E Instruction Manual  
6.9. SETUP PASS: Password Feature  
The M360E provides password protection of the calibration and setup functions to prevent unauthorized  
adjustments. When the passwords have been enabled in the PASS menu item, the system will prompt the user  
for a password anytime a password-protected function is requested.  
There are three levels of password protection, which correspond to operator, maintenance, and configuration  
functions. Each level allows access to all of the functions in the previous level.  
Table 6-9: Password Levels  
PASSWORD  
No password  
101  
LEVEL  
MENU ACCESS ALLOWED  
TEST, MSG, CLR  
Operator  
Maintenance  
Configuration  
CAL, CALZ, CALS  
818  
SETUP, VARS, DIAG  
To enable or disable passwords, press the following keystroke sequence:  
Example: If all passwords are enabled, the following keypad sequence would be required to enter the SETUP  
menu:  
SAMPLE  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
prompts for password  
number  
See Table 6-8 for  
Passwords and Levels  
SAMPLE  
ENTER SETUP PASS: 0  
ENTR EXIT  
0
0
0
8
Example: this  
password enables the  
SETUP mode  
SAMPLE  
ENTER SETUP PASS: 0  
Press individual  
keys to set  
numbers  
8
1
ENTR EXIT  
SETUP X.X  
PRIMARY SETUP MENU  
CFG DAS RNGE PASS CLK MORE  
EXIT  
Note that the instrument still prompts for a password when entering the VARS and DIAG menus, even if  
passwords are disabled, but it displays the default password (818) upon entering these menus. The user only  
has to press ENTR to access the password-protected menus but does not have to enter the required number  
code.  
84  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Model 360E Instruction Manual  
Operating Instructions  
6.10. SETUP CLK: Setting the Internal Time-of-Day Clock  
The M360E has a time of day clock that supports the AutoCal timer, time of day TEST function, and time stamps  
on most COM port messages. To set the time-of-day, press:  
SAMPLE  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
SETUP X.X  
PRIMARY SETUP MENU  
CFG DAS RNGE PASS CLK MORE  
EXIT  
SETUP X.X  
TIME-OF-DAY CLOCK  
Enter Current  
Time-of-Day  
Enter Current  
Date-of-Year  
TIME DATE  
EXIT  
SETUP X.X  
DATE: 01-JAN-02  
SETUP X.X  
TIME: 12:00  
0
1
JAN  
0
2
ENTR EXIT  
1
2
: 0  
0
ENTR EXIT  
SETUP X.X  
JAN  
DATE: 01-JAN-02  
SETUP X.X3  
: 0  
TIME: 12:00  
0
1
0
2
ENTR EXIT  
1
2
0
ENTR EXIT  
SETUP X.X  
TIME-OF-DAY CLOCK  
TIME DATE  
SETUP X.X  
EXIT  
PRIMARY SETUP MENU  
EXIT returns  
to the main  
SAMPLE display  
CFG DAS RNGE PASS CLK MORE  
EXIT  
In order to compensate for CPU clocks which run faster or slower, you can adjust a variable called CLOCK_ADJ  
to speed up or slow down the clock by a fixed amount every day. To change this variable, press:  
SAMPLE  
RANGE = 500.000 PPM  
CO2 =X.XXX  
SETUPX.X  
1 ) CONC_PRECISION = 3  
EDIT PRNT EXIT  
< TST TST > CAL  
SETUP  
PREV NEXT JUMP  
SETUP X.X  
PRIMARY SETUP MENU  
Continue to press NEXT until …  
CFG DAS RNGE PASS CLK MORE  
EXIT  
SETUP X.X  
4) CLOCK_ADJ=0 Sec/Day  
JUMP EDIT PRNT EXIT  
SETUP X.X  
SECONDARY SETUP MENU  
PREV  
COMM VARS DIAG  
EXIT  
SETUP X.X  
CLOCK_ADJ:0 Sec/Day  
ENTR EXIT  
SETUP X.X  
0 ) DAS_HOLD_OFF=15.0 Minutes  
+
0
0
NEXT JUMP  
EDIT PRNT EXIT  
Enter sign and number of seconds per  
day the clock gains (-) or loses (+).  
SETUP X.X  
4) CLOCK_ADJ=0 Sec/Day  
PREV NEXT JUMP  
EDIT PRNT EXIT  
3x EXIT returns  
to the main SAMPLE display  
05232 Rev B3  
85  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Operating Instructions  
Model 360E Instruction Manual  
6.11. SETUP MORECOMM: Using the Analyser’s  
Communication Ports  
The M360E is equipped with two serial communication ports located on the rear panel (Figure 3-2). Both ports  
operate similarly and give the user the ability to communicate with, issue commands to, and receive data from  
the analyzer through an external computer system or terminal. By default, both ports operate on the RS-232  
protocol.  
The COM1 port can also be configured to operate in single or RS-232 multidrop mode (option 62; See  
Section 5.5.2 and 6.11.7).  
The COM2 port can be configured for standard RS-232 operation, half-duplex RS-485 communication or  
for access via an LAN by installing the Teledyne Instruments Ethernet interface card (option 63; See  
Section 5.5.3 and 6.11.6).  
A code-activated switch (CAS), can also be used on either port to connect typically between 2 and 16  
send/receive instruments (host computer(s) printers, data loggers, analyzers, monitors, calibrators, etc.) into one  
communications hub. Contact Teledyne Instruments sales for more information on CAS systems.  
6.11.1. Analyzer ID  
Each type of Teledyne Instruments analyzer is configured with a default ID code. The default ID code for all  
M360M analyzers is 360. The ID number is only important if more than one analyzer is connected to the same  
communications channel such as when several analyzers are on the same Ethernet LAN (See Section 6.11.6);  
in a RS-232 multidrop chain (See Section 6.11.7) or operating over a RS-485 network (See Section 6.11.3). If  
two analyzers of the same model type are used on one channel, the ID codes of one or both of the instruments  
needs to be changed so  
To edit the instrument’s ID code, press:  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
SETUP X.X  
PRIMARY SETUP MENU  
SETUP X.X  
COMMUNICATIONS MENU  
CFG DAS RNGE PASS CLK MORE  
EXIT  
ID INET COM1  
EXIT  
Toggle these keys to  
cycle through the  
available character set:  
0-9  
ENTR key accepts the  
SETUP X.  
MACHINE ID: 360 ID  
new settings  
EXIT key ignores the new  
0
2
0
0
ENTR EXIT  
settings  
The ID number is only important if more than one analyzer is connected to the same communications channel  
(e.g., a multi-drop setup). Different models of Teledyne Instruments analyzers have different default ID  
numbers, but if two analyzers of the same model type are used on one channel (for example, two M360E’s), the  
ID of one instrument needs to be changed.  
The ID can also be used for to identify any one of several analyzers attached to the same network but situated in  
different physical locations.  
86  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Model 360E Instruction Manual  
Operating Instructions  
6.11.2. COMM Port Default Settings  
Received from the factory, the analyzer is set up to emulate a DCE or modem, with pin 3 of the DB-9 connector  
designated for receiving data and pin 2 designated for sending data.  
COM1: RS-232 (fixed) DB-9 male connector.  
o
o
o
Baud rate: 19200 bits per second (baud).  
Data Bits: 8 data bits with 1 stop bit.  
Parity: None.  
COM2: RS-232 (configurable), DB-9 female connector.  
o
o
o
Baud rate: 115000 bits per second (baud).  
Data Bits: 8 data bits with 1 stop bit.  
Parity: None.  
NOTE  
Cables that appear to be compatible because of matching connectors may incorporate  
internal wiring that make the link inoperable. Check cables acquired from sources other  
than Teledyne Instruments for pin assignments before using.  
In its default configuration, the M360E analyzer has two available RS-232 Com ports accessible via 2 DB-9  
connectors on the back panel of the instrument. The COM1 connector is a male DB-9 connector and the COM2  
is a female DB9 connector.  
Female DB-9 (COM2)  
(As seen from outside analyzer)  
Male DB-9 (RS-232)  
(As seen from outside analyzer)  
TXD  
TXD  
GND  
GND  
RXD  
RXD  
1
2
3
4
5
1
2
3
4
5
6
7
8
9
6
7
8
9
CTS  
CTS  
RTS  
RTS  
(DTE mode)  
(DTE mode)  
RXD  
GND  
TXD  
1
2
3
4
5
6
7
8
9
RTS  
CTS  
(DCE mode)  
Figure 6-8: Back Panel connector Pin-Outs for COM1 & COM2 in RS-232 mode.  
05232 Rev B3  
87  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Operating Instructions  
Model 360E Instruction Manual  
The signals from these two connectors are routed from the motherboard via a wiring harness to two 10-pin  
connectors on the CPU card, CN3 (COM1) and CN4 (COM2).  
CN3 & CN4  
(Located on CPU card)  
CTS  
RTS  
10  
RXD  
2
4
3
6
5
8
7
1
9
GND  
TXD  
(As seen from inside analyzer)  
Figure 6-9: CPU connector Pin-Outs for COM1 & COM2 in RS-232 mode.  
Teledyne Instruments offers two mating cables, one of which should be applicable for your use.  
Part number WR000077, a DB-9 female to DB-9 female cable, 6 feet long. Allows connection of COM1  
with the serial port of most personal computers. Also available as Option 60 (See Section 5.5.1).  
Part number WR000024, a DB-9 female to DB-25 male cable. Allows connection to the most common  
styles of modems (e.g. Hayes-compatible) and code activated switches.  
Both cables are configured with straight-through wiring and should require no additional adapters.  
To assist in properly connecting the serial ports to either a computer or a modem, there are activity indicators  
just above the RS-232 port. Once a cable is connected between the analyzer and a computer or modem, both  
the red and green LEDs should be on. If the lights for COM 1 are not lit, use small switch on the rear panel to  
switch it between DTE and DCE modes (See  
Section 6.11.4). If both LEDs are still not illuminated, check the cable for proper wiring.  
88  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 360E Instruction Manual  
Operating Instructions  
6.11.3. RS-485 Configuration of COM2  
As delivered from the factory, COM2 is configured for RS-232 communications. This port can be re-configured  
for operation as a non-isolated, half-duplex RS-485 port capable of supporting up to 32 instruments with a  
maximum distance between the host and the furthest instrument being 4000 feet. If you require full-duplex or  
isolated operation, please contact Teledyne Instruments Customer Service.  
To reconfigure COM2 as an RS-285 port set switch 6 of SW1 to the ON position (see Figure 6-10).  
The RS-485 port can be configured with or without a 150 termination resistor. To include the resistor,  
install jumper at position JP3 on the CPU board (see Figure 6-10). To configure COM2 as an un-  
terminated RS-485 port leave JP3 open.  
CN4  
COM2 – RS-232  
CN3  
COM1 – RS-232  
JP3  
CN5  
COM2 – RS-485  
SW1  
Pin 6  
Figure 6-10: CPU card Locations of RS-232/486 Switches, Connectors and Jumpers  
05232 Rev B3  
89  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Operating Instructions  
Model 360E Instruction Manual  
When COM2 is configured for RS-485 operation the port uses the same female DB-9 connector on the back of  
the instrument as when Com2 is configured for RS-232 operation, however, the pin assignments are different.  
Female DB-9 (COM2)  
(As seen from outside analyzer)  
RX/TX-  
GND  
RX/TX+  
1
2
3
4
5
6
7
8
9
(RS-485)  
Figure 6-11: Back Panel connector Pin-Outs for COM2 in RS-485 mode.  
The signal from this connector is routed from the motherboard via a wiring harness to a 6-pin connector on the  
CPU card, CN5.  
CN5  
(Located on CPU card)  
RX/TX  
-
GND  
RX/TX+  
2
1
4
3
6
5
(As seen from inside analyzer)  
Figure 6-12: CPU connector Pin-Outs for COM2 in RS-485 mode.  
6.11.4. DTE and DCE Communication  
RS-232 was developed for allowing communications between data terminal equipment (DTE) and data  
communication equipment (DCE). Basic terminals always fall into the DTE category whereas modems are  
always considered DCE devices. The difference between the two is the pin assignment of the Data Receive and  
Data Transmit functions. DTE devices receive data on pin 2 and transmit data on pin 3; DCE devices receive  
data on pin 3 and transmit data on pin 2.  
To allow the analyzer to be used with terminals (DTE), modems (DCE) and computers (which can be either), a  
switch mounted below the serial ports on the rear panel allows the user to switch between the two functions.  
90  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Model 360E Instruction Manual  
Operating Instructions  
6.11.5. COMM Port Communication Modes  
Each of the analyzer’s serial ports can be configured to operate in a number of different modes, listed in Table 6-  
10 which can be combined by adding the mode ID numbers. For example, quiet mode, computer mode and  
internet-enabled mode would carry a combined mode ID of 11, the standard configuration on the M360E COM2  
port. Note that each COM port needs to be configured independently.  
Table 6-10: COMM Port Communication modes  
MODE1  
QUIET  
ID  
1
DESCRIPTION  
Quiet mode suppresses any feedback from the analyzer (iDAS reports, and warning  
messages) to the remote device and is typically used when the port is communicating  
with a computer program such as APICOM. Such feedback is still available but a  
command must be issued to receive them.  
COMPUTER  
SECURITY  
Computer mode inhibits echoing of typed characters and is used when the port is  
communicating with a computer program, such as APICOM.  
2
4
When enabled, the serial port requires a password before it will respond. The only  
command that is active is the help screen (? CR).  
HESSEN  
PROTOCOL  
The Hessen communications protocol is used in some European countries. Teledyne  
Instruments part number 02252 contains more information on this protocol.  
16  
E, 7, 1  
When turned on this mode switches the COMM port settings  
from  
No parity; 8 data bits; 1 stop bit  
to  
2048  
Even parity; 7 data bits; 1 stop bit  
RS-485  
Configures the COM2 Port for RS-485 communication. RS-485 mode has precedence  
over multidrop mode if both are enabled.  
1024  
32  
MULTIDROP  
PROTOCOL  
Multidrop protocol allows a multi-instrument configuration on a single communications  
channel. Multidrop requires the use of instrument IDs.  
ENABLE  
MODEM  
Enables to send a modem initialization string at power-up. Asserts certain lines in the  
RS-232 port to enable the modem to communicate.  
64  
ERROR  
Fixes certain types of parity errors at certain Hessen protocol installations.  
CHECKING2  
128  
256  
XON/XOFF  
Disables XON/XOFF data flow control also known as software handshaking.  
HANDSHAKE2  
HARDWARE  
HANDSHAKE  
Enables CTS/RTS style hardwired transmission handshaking. This style of data  
transmission handshaking is commonly used with modems or terminal emulation  
protocols as well as by Teledyne Instrument’s APICOM software.  
8
HARDWARE  
FIFO2  
Improves data transfer rate when on of the COMM ports.  
512  
COMMAND  
PROMPT  
Enables a command prompt when in terminal mode.  
4096  
1 Modes are listed in the order in which they appear in the  
SETUP MORE COMM COM[1 OR 2] MODE menu  
2 The default sting for this feature is ON. Do not disable unless instructed to by Teledyne Instruments Customer Service  
personnel.  
05232 Rev B3  
91  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Operating Instructions  
Model 360E Instruction Manual  
Press the following keys to select a communication mode for a one of the COMM Ports, such as the following  
example where HESSEN PROTOCOL mode is enabled:  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
SETUP X.X  
PRIMARY SETUP MENU  
CFG DAS RNGE PASS CLK MORE  
EXIT  
EXIT returns to  
the previous  
menu  
SETUP X.X SECONDARY SETUP MENU  
COMM VARS DIAG ALRM  
EXIT  
EXIT  
EXIT  
SETUP X.X  
COMMUNICATIONS MENU  
Select which COM port  
to configure  
ID COM1 COM2  
The sum of the mode  
IDs of the selected  
modes is displayed here  
SETUP X.X  
SET> EDIT  
COM1 MODE:0  
SETUP X.X  
COM1 QUIET MODE: OFF  
ENTR EXIT  
NEXT OFF  
Continue pressing next until …  
SETUP X.X COM1 HESSEN PROTOCOL : OFF  
Use PREV and NEXT keys  
to move between available  
modes.  
PREV NEXT OFF  
ENTR EXIT  
A mode is enabled by  
toggling the ON/OFF key.  
ENTR key accepts the  
SETUP X.X COM1 HESSEN PROTOCOL : ON  
new settings  
EXIT key ignores the new  
PREV NEXT ON  
ENTR EXIT  
settings  
Continue pressing the NEXT and PREV keys to select any other  
modes you which to enable or disable  
92  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual  
Operating Instructions  
6.11.6. Ethernet Card Configuration  
When equipped with the optional Ethernet interface, the analyzer can be connected to any standard 10BaseT  
Ethernet network via low-cost network hubs, switches or routers. The interface operates as a standard TCP/IP  
device on port 3000. This allows a remote computer to connect through the internet to the analyzer using  
APICOM, terminal emulators or other programs.  
The firmware on board the Ethernet card automatically sets the communication modes and baud rate (115 200  
kBaud) for the COM2 port. Once the Ethernet option is installed and activated, the COM2 submenu is replaced  
by a new submenu, INET. This submenu is used to manage and configure the Ethernet interface with your LAN  
or Internet Server(s).  
The card has four LEDs that are visible on the rear panel of the analyzer, indicating its current operating status.  
Table 6-11: Ethernet Status Indicators  
LED  
LNK (green)  
ACT (yellow)  
TxD (green)  
RxD (yellow)  
FUNCTION  
ON when connection to the LAN is valid.  
Flickers on any activity on the LAN.  
Flickers when the RS-232 port is transmitting data.  
Flickers when the RS-232 port is receiving data.  
6.11.6.1. Ethernet Card COM2 Communication Modes and Baud Rate  
The firmware on board the Ethernet card automatically sets the communication modes for the COM2 port. The  
baud rate is also automatically set at 115 200 kBaud.  
6.11.6.2. Configuring the Ethernet Interface Option using DHCP  
The Ethernet option for you M360E uses Dynamic Host Configuration Protocol (DHCP) to automatically  
configure its interface with your LAN. This requires your network servers also be running DHCP. The analyzer  
will do this the first time you turn the instrument on after it has been physically connected to your network. Once  
the instrument is connected and turned on it will appear as an active device on your network without any extra  
set up steps or lengthy procedures.  
Should you need to, the Ethernet configuration properties are viewable via the analyzer’s front panel See Table  
6-12.  
05232 Rev B3  
93  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Operating Instructions  
Model 360E Instruction Manual  
Table 6-12: LAN/Internet Configuration Properties  
PROPERTY  
DEFAULT STATE  
On Editable  
DESCRIPTION  
This displays whether the DHCP is turned ON or OFF.  
DHCP STATUS  
EDIT key  
disabled when  
DHCP is ON  
INSTRUMENT  
IP ADDRESS  
Configured by  
DHCP  
This string of four packets of 1 to 3 numbers each (e.g.  
192.168.76.55.) is the address of the analyzer itself.  
EDIT key  
A string of numbers very similar to the Instrument IP  
GATEWAY IP  
ADDRESS  
Configured by  
DHCP  
disabled when address (e.g. 192.168.76.1.) that is the address of the  
DHCP is ON  
computer used by your LAN to access the Internet.  
Also a string of four packets of 1 to 3 numbers each (e.g.  
255.255.252.0) that defines that identifies the LAN the  
device is connected to.  
EDIT key  
disabled when  
DHCP is ON  
Configured by  
DHCP  
All addressable devices and computers on a LAN must  
have the same subnet mask. Any transmissions sent  
devices with different assumed to be outside of the LAN  
and are routed through gateway computer onto the  
Internet.  
SUBNET MASK  
This number defines the terminal control port by which  
the instrument is addressed by terminal emulation  
software, such as Internet or Teledyne Instruments’  
APICOM.  
TCP PORT  
3000  
Editable  
Editable  
The name by which your analyzer will appear when  
addressed from other computers on the LAN or via the  
Internet. While the default setting for all Teledyne  
Instruments M360E analyzers is “m360e” the host name  
may be changed to fit customer needs.  
HOST NAME  
M360E  
1 Do not change the setting for this property unless instructed to by Teledyne Instruments Customer Service  
personnel.  
NOTE  
It is a good idea to check these settings the first time you power up your analyzer after  
it has been physically connected to the LAN/Internet to make sure that the DHCP has  
successfully downloaded the appropriate information from you network server(s).  
If the gateway IP, instrument IP and the subnet mask are all zeroes (e.g. “0.0.0.0”),  
the DCHP was not successful.  
You may have to manually configure the analyzer’s Ethernet properties.  
See your network administrator.  
94  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 360E Instruction Manual  
Operating Instructions  
To view the above properties, press:  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
SETUP X.X  
DHCP: ON  
< TST TST > CAL  
SETUP  
SET> EDIT  
EXIT  
EXIT  
SETUP X.X  
INST IP: 0.0.0.0  
SETUP X.X  
PRIMARY SETUP MENU  
<SET SET>  
CFG DAS RNGE PASS CLK MORE  
EXIT  
EXIT  
SETUP X.X GATEWAY IP: 0.0.0.0  
SETUP X.X SECONDARY SETUP MENU  
EDIT Key  
Disabled  
<SET SET>  
EXIT  
COMM VARS DIAG  
SETUP X.X SUBNET MASK: 0.0.0.0  
SETUP X.X  
ID INET  
COMMUNICATIONS MENU  
<SET SET>  
EXIT  
EXIT  
EXIT  
COM1  
EXIT  
SETUP X.X  
TCP PORT: 3000  
SAMPLE  
ENTER SETUP PASS : 818  
8
<SET SET> EDIT  
8
1
ENTR EXIT  
SETUP X.X HOSTNAME: M200EH  
From this point on,  
EXIT returns to  
COMMUNICATIONS  
MENU  
<SET  
EDIT  
Do not alter unless  
directed to by Teledyne  
Instruments Customer  
Service personnel  
05232 Rev B3  
95  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operating Instructions  
Model 360E Instruction Manual  
6.11.6.3. Manually Configuring the Network IP Addresses  
There are several circumstances when you may need to manually configure the interface settings of the  
analyzer’s Ethernet card. The INET sub-menu may also be used to edit the Ethernet card’s configuration  
properties  
Your LAN is not running a DHCP software package,  
The DHCP software is unable to initialize the analyzer’s interface;  
You wish to program the interface with a specific set of IP addresses that may not be the ones  
automatically chosen by DHCP.  
Editing the Ethernet Interface properties is a two step process.  
STEP 1: Turn DHCP OFF: While DHCP is turned ON, the ability to manually set INSTRUMENT IP, GATEWAY  
IP and SUBNET MASK is disabled  
SAMPLE  
ENTER SETUP PASS : 818  
8
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
8
1
ENTR EXIT  
< TST TST > CAL  
SETUP  
SETUP X.X  
DHCP: ON  
SETUP X.X  
PRIMARY SETUP MENU  
<SET SET> EDIT  
EXIT  
ENTR EXIT  
ENTR EXIT  
CFG DAS RNGE PASS CLK MORE  
EXIT  
EXIT  
SETUP X.X  
DHCP: ON  
SETUP X.X SECONDARY SETUP MENU  
ON  
COMM VARS DIAG  
SETUP X.X  
DHCP: ON  
SETUP X.X  
ID INET  
COMMUNICATIONS MENU  
COM1  
OFF  
EXIT  
ENTR accept  
new settings  
Continue with editing of Ethernet interface  
properties (see Step 2, below).  
EXIT ignores  
new settings  
96  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 360E Instruction Manual  
Operating Instructions  
STEP 2: Configure the INSTRUMENT IP, GATEWAY IP and SUBNET MASK addresses by pressing:  
Internet Configuration Keypad Functions  
From Step 1 above)  
KEY  
[0]  
FUNCTION  
Press this key to cycle through the range of  
numerals and available characters (“0 – 9” & “ . ”)  
<CH CH>  
DEL  
Moves the cursor one character left or right.  
Deletes a character at the cursor location.  
SETUP X.X  
DHCP: OFF  
Accepts the new setting and returns to the previous  
menu.  
ENTR  
EXIT  
SET> EDIT  
EXIT  
EXIT  
Ignores the new setting and returns to the previous  
menu.  
Some keys only appear as needed.  
SETUP X.X INST IP: 000.000.000.000  
<SET SET> EDIT  
Cursor  
location is  
indicated by  
brackets  
SETUP X.X INST IP: [0] 00.000.000  
<CH CH>  
DEL [0]  
ENTR EXIT  
SETUP X.X GATEWAY IP: 000.000.000.000  
<SET SET> EDIT  
EXIT  
SETUP X.X GATEWAY IP: [0] 00.000.000  
<CH CH> DEL [?] ENTR EXIT  
SETUP X.X SUBNET MASK:255.255.255.0  
<SET SET> EDIT  
EXIT  
SETUP X.X SUBNET MASK:[2]55.255.255.0  
<CH CH> DEL [?] ENTR EXIT  
SETUP X.X TCP PORT 3000  
<SET  
EDIT  
EXIT  
The PORT number needs to remain at 3000.  
Do not change this setting unless instructed to by  
Teledyne Instruments Customer Service personnel.  
Pressing EXIT from  
any of the above  
display menus  
causes the Ethernet  
option to reinitialize  
its internal interface  
firmware  
SETUP X.X  
INITIALIZING INET 0%  
INITIALIZING INET 100%  
SETUP X.X  
INITIALIZATI0N SUCCEEDED  
SETUP X.X  
INITIALIZATION FAILED  
Contact your IT  
Network Administrator  
SETUP X.X  
ID INET  
COMMUNICATIONS MENU  
COM1  
EXIT  
05232 Rev B3  
97  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operating Instructions  
Model 360E Instruction Manual  
6.11.6.4. Changing the Analyzer’s HOSTNAME  
The HOSTNAME is the name by which the analyzer appears on your network. The default name for all  
Teledyne Instruments Model 360E analyzers is M360E. To change this name (particularly if you have more than  
one Model 360E analyzer on your network), press.  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
SETUP X.X  
DHCP: ON  
< TST TST > CAL  
SETUP  
SET> EDIT  
EXIT  
Continue pressing SET> UNTIL …  
SETUP X.X  
PRIMARY SETUP MENU  
CFG DAS RNGE PASS CLK MORE  
EXIT  
EXIT  
SETUP X.X HOSTNAME: 360E  
<SET EDIT  
SETUP X.X SECONDARY SETUP MENU  
EXIT  
COMM VARS DIAG ALRM  
SETUP X.X HOSTNAME: [M]360E  
SETUP X.X  
ID INET  
COMMUNICATIONS MENU  
<CH CH> INS DEL [?]  
ENTR EXIT  
COM1  
EXIT  
Use these keys (See Table 6-19)  
to edit HOSTNAME  
SAMPLE  
ENTER SETUP PASS : 818  
8
SETUP X.X HOSTNAME: 360E-FIELD1  
8
1
ENTR EXIT  
<SET  
EDIT  
EXIT  
SETUP X.X  
INITIALIZING INET 0%  
INITIALIZING INET 100%  
SETUP X.X  
INITIALIZATI0N SUCCEEDED  
SETUP X.X  
INITIALIZATION FAILED  
SETUP X.X  
ID INET  
COMMUNICATIONS MENU  
Contact your IT Network  
Administrator  
COM1  
EXIT  
Table 6-13: Internet Configuration Keypad Functions  
FUNCTION  
KEY  
<CH  
CH>  
INS  
DEL  
[?]  
Moves the cursor one character to the left.  
Moves the cursor one character to the right.  
Inserts a character before the cursor location.  
Deletes a character at the cursor location.  
Press this key to cycle through the range of numerals and characters available for insertion.  
0-9, A-Z, space ’ ~ ! # $ % ^ & * ( ) - _ = +[ ] { } < >\ | ; : , . / ?  
ENTR  
EXIT  
Accepts the new setting and returns to the previous menu.  
Ignores the new setting and returns to the previous menu.  
Some keys only appear as needed.  
98  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Model 360E Instruction Manual  
Operating Instructions  
6.11.7. Multidrop RS-232 Set Up  
The RS-232 multidrop consists of a printed circuit assembly that plugs onto the CN3, CN4, and CN5 connectors  
of the CPU card (see Figures 6-10 and 6-13) and the cabling to connect it to the analyzer’s motherboard. This  
PCA includes all circuitry required to enable your analyzer for multidrop operation. It converts the instrument’s  
COM1 port to multidrop configuration allowing up to eight analyzers to be connected the same I/O port of the  
host computer.  
Because both of the DB9 connectors on the analyzer’s back panel are needed to construct the multidrop chain,  
COM2 is no longer available for separate RS-232 or RS-485 operation; however, with the addition of an Ethernet  
Option (option 63, See Section 5.5.3 and 6.11.6) the COM2 port is available for communication over a 10BaseT  
LAN.  
JP2  
CPU Card  
Rear Panel  
(as seen from inside)  
Cable to  
Ethernet  
Card  
Multidrop  
PCA  
Cable to  
Motherboard  
Figure 6-13: Location of JP2 on RS232-Multidrop PCA (option 62)  
Each analyzer in the multidrop chain must have:  
One Teledyne Instruments option 62 installed.  
One 6’ straight-through, DB9 male DB9 Female cable (Teledyne Instruments P/N WR0000101) is  
required for each analyzer.  
To set up the network, for each analyzer:  
4. Turn the analyzer on and change its ID code (See Section 6.11.1) to a unique 4-digit number.  
5. Remove the top cover (See Section 3.1) of the analyzer and locate JP2 on the multidrop PCA (see  
Figure 6-13)  
6. Make sure that the jumpers are in place connection pins 9 10 and 11 12.  
7. If the analyzer is to be the last instrument on the chain, make sure a jumper is in place connecting pins  
21 22.  
05232 Rev B3  
99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Operating Instructions  
Model 360E Instruction Manual  
8. If you are adding an analyzer to the end of an already existing chain, don’t forget to remove JP2, pins 21  
22 on the multidrop PCA on the analyzer that was previous the last instrument in the chain.  
9. Close the instrument.  
10. Using straight-through, DB9 male DB9 Female cables interconnect the host and the analyzers as  
shown in Figure 6-14.  
NOTE:  
Teledyne Instruments recommends setting up the first link, between the Host and the  
first analyzer and testing it before setting up the rest of the chain.  
KEY:  
Host  
Female DB9  
RS-232 port  
Male DB9  
Analyzer  
Analyzer  
Analyzer  
Last Analyzer  
COM2  
COM2  
COM2  
COM2  
RS-232  
RS-232  
RS-232  
RS-232  
Make Sure  
Jumper between  
JP2 pins 21 22  
is installed.  
Figure 6-14: RS232-Multidrop PCA Host/Analyzer Interconnect Diagram  
100  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 360E Instruction Manual  
Operating Instructions  
6.11.8. COM Port Baud Rate  
To select the baud rate of one of the COM Ports, press:  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
SETUP X.X  
PRIMARY SETUP MENU  
CFG DAS RNGE PASS CLK MORE  
EXIT  
EXIT  
EXIT returns to  
the previous  
menu  
SETUP X.X SECONDARY SETUP MENU  
COMM VARS DIAG ALRM  
SETUP X.X  
COMMUNICATIONS MENU  
Select which COM port  
to configure.  
ID COM1 COM2  
EXIT  
EXIT  
EXIT  
EXIT  
EXIT  
SETUP X.X  
SET> EDIT  
COM1 MODE:0  
Press SET> until you  
reach COM1 BAUD  
RATE  
EXAMPLE  
SETUP X.X  
COM1 BAUD RATE:19200  
Use PREV and NEXT  
keys to move between  
available baud rates.  
EXIT key  
ignores the  
new setting  
<SET SET> EDIT  
300  
1200  
4800  
SETUP X.X  
COM1 BAUD RATE:19200  
ENTR  
9600  
ENTR key  
accepts the  
new setting  
19200  
38400  
57600  
115200  
PREV NEXT  
SETUP X.X  
COM1 BAUD RATE:9600  
ENTR  
NEXT ON  
05232 Rev B3  
101  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Operating Instructions  
Model 360E Instruction Manual  
6.11.9. COM Port Testing  
The serial ports can be tested for correct connection and output in the COMM menu. This test sends a string of  
256 ‘w’ characters to the selected COM port. While the test is running, the red LED on the rear panel of the  
analyzer should flicker.  
To initiate the test press the following key sequence.  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
SETUP X.X  
PRIMARY SETUP MENU  
CFG DAS RNGE PASS CLK MORE  
EXIT  
EXIT  
SETUP X.X SECONDARY SETUP MENU  
COMM VARS DIAG  
SETUP X.X  
ID COM2  
COMMUNICATIONS MENU  
COM1  
Select which  
COM port to test.  
EXIT  
EXIT  
EXIT  
SETUP X.X  
COM1 MODE:0  
SET> EDIT  
SETUP X.X  
COM1 BAUD RATE:19200  
<SET SET> EDIT  
SETUP X.X  
<SET  
COM1 : TEST PORT  
TEST  
EXIT  
SETUP X.X  
<SET  
TRANSMITTING TO COM1  
TEST  
EXIT returns to  
COMM menu  
EXIT  
Test runs  
automatically  
102  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 360E Instruction Manual  
Operating Instructions  
6.12. SETUP MORE VARS: Internal Variables (VARS)  
The M360E has several-user adjustable software variables, which define certain operational parameters.  
Usually, these variables are automatically set by the instrument’s firmware, but can be manually re-defined using  
the VARS menu. Table 6-14 lists all variables that are available within the 818 password protected level.  
Table 6-14: Variable Names (VARS) Revision B.3  
ALLOWED  
NO.  
VARIABLE  
DESCRIPTION  
VALUES  
Changes the internal data acquisition system (iDAS) hold-off  
time, which is the duration when data are not stored in the  
iDAS because the software considers the data to be  
questionable. That is the case during warm-up or just after  
the instrument returns from one of its calibration modes to  
SAMPLE mode. DAS_HOLD_OFF can be disabled entirely  
in each iDAS channel.  
Can be between 0.5  
and 20 minutes  
0
DAS_HOLD_OFF  
Default=15 min.  
Allows the user to set the number of significant digits to the  
CONC_PRECISION right of the decimal point display of concentration and  
AUTO, 1, 2, 3, 4  
Default=AUTO  
1
stability values.  
Selects which gas measurement is displayed when the STABIL  
test function is selected.  
1
STABIL_GAS  
CO; CO2 & O2  
2
3
Dynamic zero automatically adjusts offset and slope of the  
CO2 response when performing a zero point calibration  
during an AutoCal (Chapter 7).  
DYN_ZERO  
DYN_SPAN  
CLOCK_ADJ  
ON/OFF  
Dynamic span automatically adjusts slope and slope of the  
CO2 response when performing a zero point calibration  
during an AutoCal (Chapter 7).  
4
5
ON/OFF  
Note that the DYN_ZERO and DYN_SPAN features are not  
allowed for applications requiring EPA equivalency.  
Adjusts the speed of the analyzer’s clock. Choose the +  
sign if the clock is too slow, choose the - sign if the clock is  
too fast.  
-60 to +60 s/day  
1 O2 option is only available in analyzers with o2 sensor options installed.  
05232 Rev B3  
103  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Operating Instructions  
Model 360E Instruction Manual  
To access and navigate the VARS menu, use the following key sequence.  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
SETUP X.X  
PRIMARY SETUP MENU  
CFG DAS RNGE PASS CLK MORE  
EXIT  
SETUP X.X  
SECONDARY SETUP MENU  
COMM VARS DIAG ALRM  
EXIT  
SAMPLE  
ENTER SETUP PASS : 818  
EXIT ignores the new setting.  
ENTR accepts the new setting.  
8
1
8
ENTR EXIT  
SETUP X.X  
0 ) DAS_HOLD_OFF=15.0 Minutes  
SETUP X.X  
.0  
0)DAS_HOLD_OFF=15.0 Minutes  
NEXT JUMP  
EDIT PRNT EXIT  
1
5
ENTR EXIT  
Toggle this keys to change setting  
SETUP X.X  
1) CONC_PRECUISION : 3  
SETUP X.X  
1) CONC_PRECUISION : 3  
PREV NEXT JUMP  
EDIT PRNT EXIT  
AUTO  
0
1
2
3
4
ENTR EXIT  
SETUP X.X  
2 ) STABIL_GAS=CO2  
Toggle these keys to change setting  
PREV NEXT JUMP  
EDIT PRNT EXIT  
SETUP X.X  
2 ) STABIL GAS =CO2  
NO NO2 NOX O2  
ENTR EXIT  
Choose Gas  
3) DYN_ZERO=ON  
SETUP X.X  
3 ) DYN_ZERO=ON  
SETUP X.X  
PREV NEXT JUMP  
EDIT PRNT EXIT  
ON  
ENTR EXIT  
Toggle this keys to change setting  
SETUP X.X  
4) DYN_SPAN=ON  
PREV NEXT JUMP  
EDIT PRNT EXIT  
SETUP X.X  
4) DYN_SPAN=ON  
ENTR EXIT  
ON  
Toggle this keys to change setting  
SETUP X.X  
5) CLOCK_ADJ=0 Sec/Day  
SETUP X.X  
5) CLOCK_ADJ=0 Sec/Day  
ENTR EXIT  
PREV NEXT JUMP  
EDIT PRNT EXIT  
+
0
0
Toggle tese keys to change setting  
104  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual  
Operating Instructions  
6.13. SETUP MORE DIAG: Using the Diagnostics Functions  
A series of diagnostic tools is grouped together under the SETUPMOREDIAG menu. As these parameters  
are dependent on firmware revision (see Menu Tree A-6 in Appendix A). The individual parameters, however,  
are explained in more detail in the indicated in 6-15. These tools can be used in a variety of troubleshooting and  
diagnostic procedures and are referred to in many places of the maintenance and trouble-shooting s.  
Table 6-15: M360E Diagnostic (DIAG) Functions  
Front Panel  
DIAGNOSTIC FUNCTION AND MEANING  
Mode Indicator  
SIGNAL I/O: Allows observation of all digital and analog signals in  
the instrument. Allows certain digital signals such as valves and  
heaters to be toggled ON and OFF.  
DIAG I/O  
6.13.2  
6.13.3  
ANALOG OUTPUT: When entered, the analyzer performs an analog  
output step test. This can be used to calibrate a chart recorder or  
to test the analog output accuracy.  
DIAG AOUT  
ANALOG I/O CONFIGURATION: the signal levels of the instruments  
analog outputs may be calibrated (either individually or as a  
group). Various electronic parameters such as signal span, and  
offset are available for viewing and configuration.  
DIAG AIO  
6.13.4  
ELECTRIC TEST: The analyzer is performing an electric test. This  
test simulates IR detector signal in a known manner so that the  
proper functioning of the sync/demod board can be verified.  
DIAG OPTIC  
DIAG ELEC  
DIAG PCAL  
6.13.5  
6.13.6  
6.13.7  
DARK CALIBRATION: The analyzer is performing a dark  
calibration procedure. This procedure measures and stores the  
inherent dc offset of the sync/demod board electronics.  
PRESSURE CALIBRATION: The analyzer records the current  
output of the sample gas pressure sensor. This value is used by  
the CPU to compensate the CO2 concentration.  
FLOW CALIBRATION: This function is used to calibrate the gas flow  
output signals of sample gas and ozone supply. These settings  
are retained when exiting DIAG.  
DIAG FCAL  
DIAG TCHN  
6.13.8  
6.13.9  
TEST CHAN OUTPUT: Configures the A4 analog output channel.  
05232 Rev B3  
105  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Operating Instructions  
Model 360E Instruction Manual  
6.13.1. Accessing the Diagnostic Features  
To access the DIAG functions press the following keys:  
SAMPLE*  
RANGE = 500.00 PPM  
CO2 =X.XXX  
DIAG  
ANALOG I / O CONFIGURATION  
< TST TST > CAL  
SETUP  
PREV  
NEXT  
NEXT  
NEXT  
ENTR EXIT  
ENTR EXIT  
ENTR EXIT  
ENTR EXIT  
ENTR EXIT  
ENTR EXIT  
SETUP X.X  
PRIMARY SETUP MENU  
EXIT returns  
to the main  
SAMPLE  
display  
DIAG  
ELECTRICAL TEST  
DARK CALIBRATION  
CFG DAS RNGE PASS CLK MORE  
EXIT  
PREV  
SETUP X.X SECONDARY SETUP MENU  
EXIT returns  
to the PRIMARY  
SETUP MENU  
DIAG  
COMM VARS DIAG ALRM  
EXIT  
PREV  
SAMPLE  
ENTER SETUP PASS: 818  
8
DIAG  
PRESSURE CALIBRATION  
NEXT  
8
1
ENTR EXIT  
PREV  
DIAG  
SIGNAL I / O  
From this point  
forward, EXIT returns  
to the  
DIAG  
FLOW CALIBRATION  
TEST CHAN OUTPUT  
NEXT  
NEXT  
ENTR EXIT  
ENTR EXIT  
PREV  
NEXT  
NEXT  
SECONDARY  
SETUP MENU  
DIAG  
ANALOG OUTPUT  
DIAG  
PREV  
PREV  
6.13.2. Signal I/O  
The signal I/O diagnostic mode allows reviewing and changing the digital and analog input/output functions of  
the analyzer. See Appendix A-4 for a complete list of the parameters available for review under this menu.  
NOTE  
Any changes of signal I/O settings will remain in effect only until the signal I/O menu  
is exited. Exceptions are the ozone generator override and the flow sensor calibration,  
which remain as entered when exiting.  
106  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Model 360E Instruction Manual  
Operating Instructions  
To enter the signal I/O test mode, press:  
SAMPLE  
RANGE = 500.000 PPM  
CO2 =X.XXX  
DIAG  
SIGNAL I / O  
Use the NEXT & PREV  
keys to move between  
signal types.  
< TST TST > CAL  
SETUP  
PREV NEXT JUMP  
ENTR EXIT  
Use the JUMP key to  
go directly to a  
DIAG I / O  
Test Signals Displayed Here  
SETUP X.X  
PRIMARY SETUP MENU  
specific signal  
PREV NEXT JUMP  
PRNT EXIT  
CFG DAS RNGE PASS CLK MORE  
EXIT  
See Appendix A-4 for  
a complete list of  
EXIT returns  
to the main  
SAMPLE display  
available SIGNALS  
EXAMPLE  
DIAG I / O  
JUMP TO: 12  
SETUP X.X  
SECONDARY SETUP MENU  
EXAMPLE:  
1
2
ENTR EXIT  
Enter 12 to Jump to  
12) ST_CONC_VALID  
COMM VARS DIAG ALRM  
EXIT  
DIAG I / O  
ST_CONC_VALID = ON  
Exit to return  
to the  
SAMPLE  
ENTER SETUP PASS: 818  
DIAG menu  
PREV NEXT JUMP  
ON PRNT EXIT  
8
1
8
ENTR EXIT  
Pressing the PRNT key will send a formatted printout to the serial port and can be  
captured with a computer or other output device.  
6.13.3. Analog Output Step Test  
This test can be used to check the accuracy and proper operation of the analog outputs. The test forces all four  
analog output channels to produce signals ranging from 0% to 100% of the full scale range in 20% increments.  
This test is useful to verify the operation of the data logging/recording devices attached to the analyzer.  
To begin the Analog Output Step Test press:  
DIAG  
SIGNAL I / O  
SAMPLE  
RANGE = 500.000 PPM  
CO2 =X.XXX  
NEXT  
ENTR EXIT  
< TST TST > CAL  
SETUP  
DIAG  
ANALOG OUTPUT  
SETUP X.X  
PRIMARY SETUP MENU  
PREV  
NEXT  
ENTR EXIT  
CFG DAS RNGE PASS CLK MORE  
EXIT  
Performs  
analog output  
step test.  
DIAG AOUT  
0%  
ANALOG OUTPUT  
ANALOG OUTPUT  
SETUP X.X  
SECONDARY SETUP MENU  
0% - 100%  
EXIT  
COMM VARS DIAG ALRM  
EXIT  
Exit-Exit  
returns to the  
DIAG menu  
DIAG AOUT  
SETUP X.X  
ENTER DIAG PASS: 818  
8
1
8
ENTR EXIT  
[0%]  
EXIT  
Pressing the key under “0%” while performing the test will  
pause the test at that level. Brackets will appear around  
the value: example: [20%] Pressing the same key again  
will resume the test.  
05232 Rev B3  
107  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Operating Instructions  
Model 360E Instruction Manual  
6.13.4. Analog I/O Configuration  
The analog I/O functions that are available in the M360E are:  
Table 6-16: DIAG - Analog I/O Functions  
Function  
Sub Menu  
AOUTS CALIBRATED: Shows the status of the analog output calibration (YES/NO) and initiates a calibration of all  
analog output channels.  
CONC_OUT_1  
Sets the basic electronic configuration of the A1 analog output (CO2). There are three options:  
Range: Selects the signal type (voltage or current loop) and full scale level of the output.  
REC_OFS: Allows setting a voltage offset (not available when RANGE is set to CURRent loop.  
Auto_CAL: Performs the same calibration as AOUT CALIBRATED, but on this one channel only.  
NOTE: Any change to RANGE or REC_OFS requires recalibration of this output.  
Same as for CONC_OUT_1 but for analog channel 2 (CO2 )  
CONC_OUT_2  
TEST OUTPUT  
Same as for CONC_OUT_1 but for analog channel 4 (TEST)  
AIN CALIBRATED  
Shows the calibration status (YES/NO) and initiates a calibration of the analog to digital  
converter circuit on the motherboard.  
To configure the analyzer’s three analog outputs, set the electronic signal type of each channel and calibrate the  
outputs. This consists of:  
Selecting an output type (voltage or current, if an optional current output driver has been installed) and  
the signal level that matches the input requirements of the recording device attached to the channel, See  
Section 6.13.4.1.  
Calibrating the output channel. This can be done automatically or manually for each channel, see  
Section 6.13.4.2 and 6.13.4.3.  
Adding a bipolar recorder offset to the signal, if required (Section 6.13.4.2.)  
In its standard configuration, the analyzer’s outputs can be set for the following DC voltages. Each range is  
usable from -5% to + 5% of the nominal range.  
Table 6-17: Analog Output Voltage Ranges  
RANGE  
0-0.1 V  
0-1 V  
MINIMUM OUTPUT  
-5 mV  
MAXIMUM OUTPUT  
+105 mV  
-0.05 V  
+1.05 V  
0-5 V  
-0.25 V  
+5.25 V  
0-10 V  
-0.5 V  
+10.5 V  
The default offset for all ranges is 0 VDC.  
108  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Model 360E Instruction Manual  
Operating Instructions  
The following DC current output limits apply to the current loop modules:  
Table 6-18: Analog Output Current Loop Range  
RANGE  
MINIMUM OUTPUT  
MAXIMUM OUTPUT  
0-20 mA  
0 mA  
20 mA  
These are the physical limits of the current loop modules, typical  
applications use 2-20 or 4-20 mA for the lower and upper limits. Please  
specify desired range when ordering this option.  
The default offset for all ranges is 0 mA.  
Pin assignments for the output connector at the rear panel of the instrument are shown in Table 6-19.  
ANALOG OUT  
A1  
A2  
A3  
A4  
+
-
+
-
+
-
+
-
Table 6-19: Analog Output Pin Assignments  
PIN  
ANALOG  
OUTPUT  
VOLTAGE  
SIGNAL  
CURRENT  
SIGNAL  
1
2
V Out  
I Out +  
I Out -  
A1  
A2  
Ground  
3
4
V Out  
I Out +  
I Out -  
Ground  
5
6
V Out  
I Out +  
I Out -  
A31  
A4  
Ground  
7
8
V Out  
I Out +  
I Out -  
Ground  
1 Output A3 is only used when the O2 sensor option is installed  
See Figure 3-2 for the location of the analog output connector on the instrument’s rear panel.  
05232 Rev B3  
109  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Operating Instructions  
Model 360E Instruction Manual  
6.13.4.1. Analog Output Signal Type and Range Span Selection  
To select an output signal type (DC Voltage or current) and level for one output channel, activate the ANALOG  
I/O CONFIGURATION MENU (See Section 6.13.1) then press:  
FROM ANALOG I/O CONFIGURATION MENU  
DIAG  
ANALOG I / O CONFIGURATION  
PREV  
NEXT  
ENTR  
EXIT  
EXIT  
DIAG AIO  
AOUTS CALIBRATED: NO  
Press SET> to select the  
analog output channel to be  
configured. Press EDIT to  
continue  
< SET SET> CAL  
DIAG AIO  
CONC_OUT_2:5V, CAL  
< SET SET> EDIT  
EXIT  
EXIT  
DIAG AIO  
CONC_OUT_2 RANGE: 5V  
SET> EDIT  
DIAG AIOOUTPUT RANGE: 5V  
These keys set  
the signal level  
and type of the  
selected channel  
0.1V 1V 5V 10V CURR  
ENTR EXIT  
Pressing ENTR records the new setting  
and returns to the previous menu.  
Pressing EXIT ignores the new setting and  
returns to the previous menu.  
DIAG AIOOUTPUT RANGE: 10V  
0.1V 1V 5V 10V CURR  
ENTR EXIT  
6.13.4.2. Analog Output Calibration Mode  
The analog outputs can be calibrated automatically or manually. In its default mode, the instrument is  
configured for automatic calibration of all channels. Manual calibration should be used for the 0.1V range or in  
cases where the outputs must be closely matched to the characteristics of the recording device. Outputs  
configured for automatic calibration can be calibrated as a group or individually. Calibration of the analog  
outputs needs to be carried out on first startup of the analyzer (performed in the factory as part of the  
configuration process) or whenever re-calibration is required.  
110  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Model 360E Instruction Manual  
Operating Instructions  
To calibrate the outputs as a group, activate the ANALOG I/O CONFIGURATION MENU (See Section 6.13.1),  
then press:  
STARTING FROM DIAGNOSTIC MENU  
(see Section 6.13.1)  
DIAG  
ANALOG I / O CONFIGURATION  
ENTR EXIT  
Exit at any time  
to return to the  
main DIAG  
menu  
PREV  
NEXT  
DIAG AIO  
AOUTS CALIBRATED: NO  
If AutoCal has been  
turned off for any  
channel, the message  
for that channel will be  
similar to:  
< SET SET> CAL  
EXIT  
DIAG AIO AUTO CALIBRATING CONC_OUT_1  
AUTO CALIBRATING CONC_OUT_2  
NOT AUTO CAL  
CONC_OUT_1  
AUTO CALIBRATING TEST_OUTPUT  
If any of the channels have  
not been calibrated this  
message will read NO.  
Exit to return to  
the I/O  
configuration  
menu  
DIAG AIO  
AOUTS CALIBRATED:  
YES  
< SET SET> CAL  
EXIT  
To automatically calibrate a single analog channel, activate the ANALOG I/O CONFIGURATION MENU (See  
Section 6.13.1), then press:  
DIAG  
ANALOG I / O CONFIGURATION  
NEXT  
ENTR EXIT  
EXIT to Return  
to the main  
Sample Display  
PREV  
DIAG AIO  
AOUTS CALIBRATED: NO  
<
SET> CAL  
EXIT  
Press SET> to select the  
Analog Output channel to  
be configured. Then Press  
EDIT to continue  
DIAG AIO  
CONC_OUT_2:5V, CAL  
< SET SET> EDIT  
EXIT  
EXIT  
DIAG AIO  
CONC_OUT_2 RANGE: 5V  
DIAG AIO  
CONC_OUT_2 CALIBRATED: NO  
CAL EXIT  
SET> EDIT  
<SET  
DIAG AIO  
CONC_OUT_2 REC OFS: 0 mV  
DIAG AIO  
AUTO CALIBRATING CONC_OUT_2  
< SET SET> EDIT  
EXIT  
DIAG AIO  
CONC_OUT_2 AUTO CAL: ON  
DIAG AIO  
<SET  
CONC_OUT_2 CALIBRATED: YES  
< SET SET> EDIT  
EXIT  
CAL  
EXIT  
05232 Rev B3  
111  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operating Instructions  
Model 360E Instruction Manual  
To select manual output calibration for a particular channel, activate the ANALOG I/O CONFIGURATION MENU  
(See Section 6.13.1), then press:  
DIAG  
ANALOG I / O CONFIGURATION  
ENTR EXIT  
DIAG AIO  
CONC_OUT_2 REC OFS: 0 mV  
Exit to return to  
the main  
sample display  
PREV  
NEXT  
< SET SET> EDIT  
EXIT  
EXIT  
DIAG AIO  
AOUTS CALIBRATED: NO  
DIAG AIO  
CONC_OUT_2 AUTO CAL: ON  
< SET SET> CAL  
EXIT  
< SET SET> EDIT  
Press SET> to select the analog output channel to  
be configured. Then press EDIT to continue  
DIAG AIO  
CONC_OUT_2 AUTO CAL: ON  
DIAG AIO  
CONC_OUT_2:5V, CAL  
ON  
ENTR EXIT  
< SET SET> EDIT  
EXIT  
EXIT  
Toggles the  
auto cal mode  
ON/ OFF for  
this analog  
output channel  
only.  
ENTR accepts the new setting  
and returns to the previous  
menu. EXIT ignores the new  
setting and returns to the  
previous menu.  
DIAG AIO  
CONC_OUT_2 RANGE: 5V  
SET> EDIT  
Now the analog output channels should either be automatically calibrated or they should be set to manual  
calibration, which is described next.  
6.13.4.3. Manual Analog Output Calibration and Voltage Adjustment  
For highest accuracy, the voltages of the analog outputs can be manually calibrated. Calibration is done through  
the instrument software with a voltmeter connected across the output terminals (Figure 6-15). Adjustments are  
made using the front panel keys by setting the zero-point first and then the span-point (Table 6-20).  
The software allows this adjustment to be made in 100, 10 or 1 count increments.  
Table 6-20: Voltage Tolerances for Analog Output Calibration  
Full Scale  
0.1 VDC  
1 VDC  
Zero Tolerance  
±0.0005V  
±0.001V  
Span Voltage  
90 mV  
Span Tolerance  
±0.001V  
900 mV  
±0.001V  
5 VDC  
±0.002V  
4500 mV  
4500 mV  
±0.003V  
10 VDC  
±0.004V  
±0.006V  
NOTE  
Outputs configured for 0.1V full scale should always be calibrated manually  
112  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Model 360E Instruction Manual  
Operating Instructions  
See Table 3-1 for  
pin assignments  
of Analog Out  
connector on the  
rear panel  
V
+DC Gnd  
V OUT +  
V OUT -  
V IN +  
V IN -  
Recording  
Device  
ANALYZER  
Figure 6-15: Setup for Calibrating Analog Voltage Outputs  
To make these adjustments, the AOUT auto-calibration feature must be turned off (Section 6.13.4.2). Activate  
the ANALOG I/O CONFIGURATION MENU (See Section 6.13.1), then press:  
FROM ANALOG I/O CONFIGURATION MENU  
DIAG AIO  
CONC_OUT_1 RANGE: 5V  
DIAG  
ANALOG I / O CONFIGURATION  
ENTR EXIT  
SET> EDIT  
EXIT  
PREV  
NEXT  
DIAG AIO  
CONC_OUT_1 REC OFS: 0 mV  
DIAG AIO  
AOUTS CALIBRATED: NO  
< SET SET> EDIT  
EXIT  
EXIT  
If AutoCal is ON, go to  
Section 6.7.3  
< SET SET> CAL  
EXIT  
DIAG AIO  
CONC_OUT_1 AUTO CAL: OFF  
Press SET> to select the analog output channel to be  
< SET SET> EDIT  
configured:  
DISPLAYED AS=  
CONC_OUT_1 =  
CONC_OUT_2 =  
TEST OUTPUT =  
CHANNEL  
A1  
A2  
A4  
DIAG AIO  
< SET  
CONC_OUT_2 CALIBRATED: NO  
CAL  
EXIT  
DIAG AIO  
CONC_OUT_1 :5V, NO CAL  
DIAG AIO CONC_OUT_1 VOLT–Z : 0 mV  
< SET SET> EDIT  
EXIT  
U100 UP10 UP DOWN DN10 D100 ENTR EXIT  
These keys increase / decrease the analog  
output by 100, 10 or 1 counts.  
EXIT ignores the  
new setting.  
ENTR accepts the  
Continue adjustments until the voltage measured  
at the output of the analyzer and/or the input of  
the recording device matches the value in the  
upper right hand corner of the display to the  
tolerance listed in Table 6-20.  
DIAG AIO CONC_OUT_1 VOLT–S : 4500 mV  
U100 UP10 UP DOWN DN10 D100 ENTR EXIT  
new setting.  
The concentration display will not change. Only  
the voltage reading of your voltmeter will change.  
DIAG AIO  
< SET  
CONC_OUT_1 CALIBRATED: YES  
CAL EXIT  
05232 Rev B3  
113  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Operating Instructions  
Model 360E Instruction Manual  
6.13.4.4. Current Loop Output Adjustment  
A current loop option is available and can be installed as a retrofit for each of the analog outputs of the analyzer  
(s 5.2). This option converts the DC voltage analog output to a current signal with 0-20 mA output current. The  
outputs can be scaled to any set of limits within that 0-20 mA range. However, most current loop applications  
call for either 2-20 mA or 4-20 mA range. All current loop outputs have a +5% over-range. Ranges with the  
lower limit set to more than 1 mA (e.g., 2-20 or 4-20 mA) also have a -5% under-range.  
To switch an analog output from voltage to current loop after installing the current output printed circuit  
assembly, follow the instructions in Section 6.13.4.4 and select CURR from the list of options on the RANGE  
menu.  
Adjusting the signal zero and span values of the current loop output is done by raising or lowering the voltage of  
the respective analog output. This proportionally raises or lowers the current produced by the current loop  
option.  
Similar to the voltage calibration, the software allows this current adjustment to be made in 100, 10 or 1 count  
increments. Since the exact current increment per voltage count varies from output to output and from  
instrument to instrument, you will need to measure the change in the current with a current meter placed in  
series with the output circuit (Figure 6-16).  
See Table 3-1 for  
pin assignments of  
mA  
the Analog Out  
connector on the  
rear panel.  
IN  
OUT  
V OUT +  
V OUT -  
I IN +  
I IN -  
Recording  
Device  
Analyzer  
Figure 6-16: Setup for Calibrating Current Outputs  
NOTE  
Do not exceed 60 V between current loop outputs and instrument ground.  
To adjust the zero and span values of the current outputs, activate the ANALOG I/O CONFIGURATION MENU  
(See Section 6.13.1), then press:  
114  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Model 360E Instruction Manual  
Operating Instructions  
FROM ANALOG I/O CONFIGURATION MENU  
The instrument attempt to automatically calibrate  
the channel … then beep.  
DIAG  
ANALOG I / O CONFIGURATION  
NEXT ENTR  
PREV  
EXIT  
DIAG AIO CONC_OUT_2 D/A/ CAL ERROR  
EXIT  
DIAG AIO  
AIN CALIBRATED: NO  
EXIT  
SET> EDIT  
DIAG AIO CONC_OUT_2 CURR-Z: 0 mV  
U100 UP10 UP DOWN DN10 D100 ENTR EXIT  
Press SET> to select the analog output channel  
to be configured:.  
DIAG AIO CONC_OUT_2 ZERO: 27 mV  
Increase or decrease the current  
output by 100, 10 or 1 counts.  
The resulting change in output  
voltage is displayed in the upper  
line.  
U100 UP10 UP DOWN DN10 D100 ENTR EXIT  
DIAG AIO CONC_OUT_2:CURR, NO CAL  
< SET SET> EDIT  
EXIT  
Continue adjustments until the  
correct current is measured with  
the current meter.  
DIAG AIO CONC_OUT_2 SPAN: 10000 mV  
DIAG AIO CONC_OUT_2 RANGE: CURR  
U100 UP10 UP DOWN DN10 D100 ENTR EXIT  
<SET SET> EDIT  
EXIT  
EXIT  
DIAG AIO CONC_OUT_2 ZERO: 9731 mV  
EXIT ignores the  
new setting, ENTR  
accepts the new  
setting.  
DIAG AIO CONC_OUT_2 CALIBRATED: NO  
U100 UP10 UP DOWN DN10 D100 ENTR EXIT  
< SET  
CAL  
DIAG AIO CONC_OUT_2 CALIBRATED: YES  
DIAG AIO AUTO CALIBRATING CONC_OUT_2  
< SET  
CAL  
EXIT  
If a current meter is not available, an alternative method for calibrating the current loop outputs is to connect a  
250   1% resistor across the current loop output. Using a voltmeter connected across the resistor, follow the  
procedure above but adjust the output to the following values:  
Table 6-21: Current Loop Output Calibration with Resistor  
Voltage for 2-20 mA  
(measured across resistor)  
Voltage for 4-20 mA  
(measured across resistor)  
Full scale  
0%  
0.5 V  
5.0 V  
1.0 V  
5.0 V  
100%  
05232 Rev B3  
115  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Operating Instructions  
Model 360E Instruction Manual  
6.13.4.5. AIN Calibration  
This is the sub-menu to conduct the analog input calibration. This calibration should only be necessary after  
major repair such as a replacement of CPU, motherboard or power supplies. Activate the ANALOG I/O  
CONFIGURATION MENU (See Section 6.13.1), then press:  
STARTING FROM ANALOG I / O CONFIGURATION MENU  
Exit at any time to  
return to the main  
DIAG menu  
DIAG  
ANALOG I / O CONFIGURATION  
ENTR EXIT  
PREV  
NEXT  
Continue pressing SET? until …  
DIAG AIO  
AIN CALIBRATED: NO  
< SET SET> CAL  
EXIT  
DIAG AIO  
CALIBRATING A/D ZERO  
Instrument  
calibrates  
automatically  
CALIBRATING A/D SPAN  
Exit to return to the  
ANALOG I/O  
CONFIGURATION  
MENU  
DIAG AIO  
AIN CALIBRATED: YES  
< SET SET> CAL  
EXIT  
116  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 360E Instruction Manual  
Operating Instructions  
6.13.5. Electric Test  
The electric test function substitutes simulated signals for CO2 MEAS and CO2 REF, generated by circuitry on  
the sync/demod board, for the output of the IR photo-detector. While in this mode the user can also view the  
same test functions viewable from the main SAMPLE display. When the test is running, the concentration  
reported on the front panel display should be 40.0 ppm.  
SAMPLE  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
SETUP X.X  
PRIMARY SETUP MENU  
CFG DAS RNGE PASS CLK MORE  
EXIT  
EXIT  
SETUP X.X  
SECONDARY SETUP MENU  
COMM VARS DIAG ALRM  
SETUP X.X  
ENTER DIAG PASS: 818  
8
1
8
ENTR EXIT  
DIAG  
SIGNAL I / O  
NEXT  
ENTR  
EXIT  
Repeat Pressing NEXT unti . . .  
DIAG  
ELECTRIC TEST  
PREV NEXT  
ENTR  
EXIT  
DIAG ELEC  
RANGE=50.000 PPM  
CO2= 40.0  
EXIT  
Exit returns  
to the  
DIAG Menu  
<TST TST>  
Press <TST TST> to view Test Functions  
NOTE: CO MEAS and CO REF will be artificially altered to  
enforce a CO2 reading of 40.0 ppm.  
All other Test Functions will report the correct operational  
value  
05232 Rev B3  
117  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Operating Instructions  
Model 360E Instruction Manual  
6.13.6. Dark Calibration Test  
The dark calibration test interrupts the signal path between the IR photo-detector and the remainder of the  
sync/demod board circuitry. This allows the instrument to compensate for any voltage levels inherent in the  
sync/demod circuitry that might effect the calculation of CO2 concentration. Performing this calibration returns  
two offset voltages, One for CO2 MEAS and on for CO2 REF that are automatically added to the CPU’s  
calculation routine. The two offset voltages from the last calibration procedure may be reviewed by the user via  
the front panel display.  
To activate the dark calibration procedure or review the results of a previous calibration, press:  
SAMPLE  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
SETUP X.X  
PRIMARY SETUP MENU  
CFG DAS RNGE PASS CLK MORE  
EXIT  
EXIT  
DIAG  
SIGNAL I / O  
NEXT  
ENTR  
EXIT  
SETUP X.X  
SECONDARY SETUP MENU  
COMM VARS DIAG ALRM  
Repeat Pressing NEXT until . . .  
SETUP X.X  
ENTER DIAG PASS: 818  
DIAG  
DARK CALIBRATION  
8
1
8
ENTR EXIT  
PREV NEXT  
ENTR  
EXIT  
DIAG DARK  
CO DARK CALIBRATION  
Exit returns  
to the  
previous menu  
VIEW CAL  
EXIT  
Calibration runs automatically  
Electric offset for Reference signal  
Display  
tracks %  
complete  
DIAG DARK  
REF DARK OFFSET: 0.0 mV  
DIAG DARK  
DARK CAL 1% COMPLETE  
EXIT  
EXIT  
EXIT  
EXIT  
Electric offset for Measurement signal  
DIAG DARK  
MEAS DARK OFFSET: 0.0 mV  
DIAG DARK  
DARK CALIBRATION ABORTED  
118  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 360E Instruction Manual  
Operating Instructions  
6.13.7. Pressure Calibration  
A sensor at the exit of the sample chamber continuously measures the pressure of the sample gas. This data is  
used to compensate the final CO2 concentration calculation for changes in atmospheric pressure and is stored in  
the CPU’s memory as the test function PRES (also viewable via the front panel).  
NOTE  
This calibration must be performed when the pressure of the sample gas is equal to  
ambient atmospheric pressure.  
Before performing the following pressure calibration procedure, disconnect the sample  
gas pump and the sample gas-line vent from the sample gas inlet on the instrument’s  
rear panel.  
To cause the analyzer to measure and record a value for PRES, press.  
SAMPLE  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
SETUP X.X  
PRIMARY SETUP MENU  
Exit at  
any time  
to return  
to main  
the  
CFG DAS RNGE PASS CLK MORE  
EXIT  
EXIT  
SETUP  
menu  
SETUP X.X  
SECONDARY SETUP MENU  
COMM VARS DIAG ALRM  
SETUP X.X  
ENTER DIAG PASS: 818  
8
1
8
ENTR EXIT  
DIAG  
SIGNAL I / O  
NEXT  
ENTR  
EXIT  
Repeat Pressing NEXT until . . .  
ENTR accepts the  
new value and  
returns to the  
DIAG PCAL ACTUAL PRESS : 27.20 IN-HG-A  
previous menu  
EXIT ignores the  
new value and  
returns to the  
2
7
.2  
0
ENTR EXIT  
Adjust these values until the  
displayed pressure equals the  
pressure measured by the  
independent pressure meter.  
previous menu  
05232 Rev B3  
119  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Operating Instructions  
Model 360E Instruction Manual  
6.13.8. Flow Calibration  
The flow calibration allows the user to adjust the values of the sample flow rates as they are displayed on the  
front panel and reported through COM ports to match the actual flow rate measured at the sample inlet. This  
does not change the hardware measurement of the flow sensors, only the software calculated values.  
To carry out this adjustment, connect an external, sufficiently accurate flow meter to the sample inlet (see  
Chapter 11 for more details). Once the flow meter is attached and is measuring actual gas flow, press:  
SAMPLE  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
SETUP X.X  
PRIMARY SETUP MENU  
CFG ACAL DAS RNGE PASS CLK MORE EXIT  
Exit at  
any time  
to return  
to main  
the  
SETUP X.X  
SECONDARY SETUP MENU  
COMM VARS DIAG  
EXIT  
SETUP  
menu  
SETUP X.X  
ENTER DIAG PASS: 818  
8
1
8
ENTR EXIT  
DIAG  
SIGNAL I / O  
NEXT  
ENTR EXIT  
Adjust these values until  
the displayed flow rate  
equals the flow rate being  
measured by the  
independent flow meter.  
Repeat Pressing NEXT until . . .  
DIAG  
FLOW CALIBRATION  
Exit returns  
to the  
PREV NEXT  
ENTR EXIT  
previous menu  
DIAG FCAL  
ACTUAL FLOW: 654 CC / M  
ENTR EXIT  
Adjust these values  
until the displayed  
flow rate equals the  
flow rate being  
measured by the  
independent flow  
meter.  
ENTR accepts the  
new value and  
returns to the  
previous menu  
EXIT ignores the  
new value and  
returns to the  
0
6
5
4
previous menu  
120  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 360E Instruction Manual  
Operating Instructions  
6.13.9. Test Channel Output  
When activated, output channel A4 can be used to report one of the test functions viewable from the SAMPLE  
mode display. To activate the A4 channel and select a test function, follow this key sequence:  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
SETUP X.X  
PRIMARY SETUP MENU  
Continue to press NEXT until …  
CFG DAS RNGE PASS CLK MORE  
EXIT  
EXIT  
SETUP X.X SECONDARY SETUP MENU  
DIAG  
TEST CHAN OUTPUT  
COMM VARS DIAG ALRM  
PREV  
NEXT  
ENTR  
EXIT  
SETUP X.X  
ENTER DIAG PASS: 818  
DIAG TCHN  
TEST CHANNEL: NONE  
8
1
8
ENTR EXIT  
NEXT  
ENTR  
EXIT  
DIAG  
SIGNAL I / O  
DIAG TCHN TEST CHANNEL: CO2 MEASURE  
PREV NEXT ENTR  
NEXT  
ENTR EXIT  
ENTR EXIT  
EXIT  
DIAG  
ANALOG OUTPUT  
PREV NEXT  
Press PREV or NEXT  
to move through the  
list of available  
parameters  
Press ENTR to select  
Press EXIT to  
return to the  
DIAG menu  
the displayed  
parameter activating  
the test channel.  
(Table 6-13)  
Table 6-22: Test Parameters Available for Analog Output A4  
TEST CHANNEL  
ZERO  
FULL SCALE  
NONE  
Test Channel is turned off  
CO2 MEASURE  
CO2 REFERENCE  
SAMPLE PRESS  
SAMPLE FLOW  
SAMPLE TEMP  
BENCH TEMP  
WHEEL TEMP  
CHASSIS TEMP  
PHT DRIVE  
0 mV  
0 mV  
5000 mV*  
5000 mV*  
40 "Hg  
0 "Hg  
0 cc/m  
1000 cc/m  
0C  
0C  
70C  
70C  
0C  
70C  
0C  
70C  
0 mV  
5000 mV  
* This refers to the internal voltage level of the function NOT the output signal level  
of the Test channel itself.  
05232 Rev B3  
121  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Operating Instructions  
Model 360E Instruction Manual  
6.14. SETUP MORE ALRM: Using the Gas Concentration  
Alarms  
The Model 360E includes two CO2 concentration alarms. Each alarm has a user settable limit, and is associated  
with an opto-isolated TTL relay accessible via the status output connector on the instrument’s back panel (See  
Section 6.15.1.1). If the CO2 concentration measured by the instrument rises above that limit, the alarm‘s status  
output relay is closed.  
The default settings for ALM1 and ALM2 are:  
Table 6-23: CO2 Concentration Alarm Default Settings  
1
ALARM  
STATUS  
LIMIT SET POINT  
100 ppm  
ALM1  
ALM2  
Disabled  
Disabled  
300 ppm  
1
Set points listed are for PPM. Should the reporting range units of measure be changed (See Section 6.7.6) the  
analyzer will automatically scale the set points to match the new range unit setting.  
Note  
To prevent the concentration alarms from activating during span calibration operations  
make sure to press CAL or CALS button prior to introducing span gas into the analyzer.  
6.14.1. Setting the Concentration Alarm Limits  
To enable either of the CO2 concentration alarms and set the Limit points, press:  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
SETUP X.X  
ALARM MENU  
SETUP X.X  
PRIMARY SETUP MENU  
ALM1 ALM2  
EXIT  
CFG DAS RNGE PASS CLK MORE  
EXIT  
EXIT  
SETUP X.  
ALARM 1 LIMIT: OFF  
SETUP X.X SECONDARY SETUP MENU  
OFF  
ENTR EXIT  
COMM VARS DIAG ALRM  
SETUP X.  
ALARM 1 LIMIT: ON  
ON  
ENTR EXIT  
Toggle these keys to  
cycle through the  
available character set:  
0-9  
ENTR key accepts the  
SETUP X.  
ALARM 1 LIMIT: 200,00 PPM  
.0 ENTR EXIT  
new settings  
EXIT key ignores the new  
0
1
0
0
0
settings  
122  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Model 360E Instruction Manual  
Operating Instructions  
6.15. Remote Operation of the Analyzer  
6.15.1. Remote Operation Using the External Digital I/O  
6.15.1.1. Status Outputs  
The status outputs report analyzer conditions via optically isolated NPN transistors, which sink up to 50 mA of  
DC current. These outputs can be used interface with devices that accept logic-level digital inputs, such as  
programmable logic controllers (PLC’s). Each Status bit is an open collector output that can withstand up to 40  
VDC. All of the emitters of these transistors are tied together and available at D.  
NOTE  
Most PLC’s have internal provisions for limiting the current that the input will draw from  
an external device. When connecting to a unit that does not have this feature, an  
external dropping resistor must be used to limit the current through the transistor  
output to less than 50 mA. At 50 mA, the transistor will drop approximately 1.2V from  
its collector to emitter.  
The status outputs are accessed via a 12-pin connector on the analyzer’s rear panel labeled STATUS. The  
function of each pin is defined in Table 6–24.  
STATUS  
1
2
3
4
5
6
7
8
D
+
Figure 6-17: Status Output Connector  
05232 Rev B3  
123  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Operating Instructions  
Model 360E Instruction Manual  
The pin assignments for the Status Outputs are:  
Table 6-24: Status Output Pin Assignments  
OUTPUT #  
Status Definition  
Condition  
On if no faults are present.  
1
2
SYSTEM OK  
On if CO2 concentration measurement is valid.  
CONC VALID  
If the CO2 concentration measurement is invalid, this bit is OFF.  
On if unit is in high range of DUAL or AUTO range modes.  
On whenever the instruments zero point is being calibrated.  
On whenever the instruments span point is being calibrated.  
On whenever the instrument is in diagnostic mode.  
3
4
5
6
HIGH RANGE  
ZERO CAL  
SPAN CAL  
DIAG MODE  
On whenever the measured CO2 concentration is above the set point for  
ALM1  
7
8
ALARM1  
ALARM2  
On whenever the measured CO2 concentration is above the set point for  
ALM2  
D
+
EMITTER BUSS  
DC POWER  
The emitters of the transistors on pins 1-8 are bussed together.  
+ 5 VDC  
Digital Ground  
The ground level from the analyzer’s internal DC power supplies.  
124  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 360E Instruction Manual  
Operating Instructions  
6.15.1.2. Control Inputs  
These inputs allow the user to remotely initiate Zero and Span calibrations. Two methods for energizing the  
inputs is provided below; the first using the internal +5V available on the CONTROL IN connector and the  
second, if an external, isolated supply is employed.  
Table 6-25: Control Input Pin Assignments  
INPUT  
STATUS  
CONDITION WHEN ENABLED  
EXTERNAL ZERO  
CAL  
Zero calibration mode is activated. The mode field of the display will  
read ZERO CAL R.  
A
EXTERNAL SPAN  
CAL  
Span calibration mode is activated. The mode field of the display will  
read SPAN CAL R.  
B
C
D
E
F
Unused  
Unused  
Unused  
Unused  
DIGITAL GROUND  
Provided to ground an external device (e.g., recorder).  
DC power for Input  
pull ups  
Input for +5 VDC required to activate inputs A - F. This voltage can be  
taken from an external source or from the “+” pin.  
U
+
Internal source of +5V which can be used to activate inputs when  
connected to pin U.  
Internal +5V Supply  
There are two methods to activate control inputs. The internal +5V available from the “+” pin is the most  
convenient method (Figure 6-18). However, to ensure that these inputs are truly isolated, a separate, external 5  
VDC power supply should be used.  
CONTROL IN  
CONTROL IN  
A
B
C
D
E
F
U
+
A
B
C
D
E
F
U
+
Z
E
R
O
S
P
A
N
Z
E
R
O
S
P
A
N
5 VDC Power  
Supply  
+
-
External Power Connections  
Local Power Connections  
Figure 6-18: Control Inputs  
05232 Rev B3  
125  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Operating Instructions  
Model 360E Instruction Manual  
6.15.2. Remote Operation Using the External Serial I/O  
6.15.2.1. Terminal Operating Modes  
The Model 360E can be remotely configured, calibrated or queried for stored data through the serial ports. As  
terminals and computers use different communication schemes, the analyzer supports two communicate modes  
specifically designed to interface with these two types of devices.  
Computer mode is used when the analyzer is connected to a computer with a dedicated interface  
program such as APICOM. More information regarding APICOM can be found in later in this or on the  
Teledyne Instruments website at http://www.teledyne-api.com/software/apicom/.  
Interactive mode is used with a terminal emulation programs such as HyperTerminal or a “dumb”  
computer terminal. The commands that are used to operate the analyzer in this mode are listed in Table  
6-26.  
6.15.2.2. Help Commands in Terminal Mode  
Table 6-26: Terminal Mode Software Commands  
COMMAND  
Control-T  
Function  
Switches the analyzer to terminal mode (echo, edit). If mode flags 1 & 2 are OFF, the interface  
can be used in interactive mode with a terminal emulation program.  
Control-C  
Switches the analyzer to computer mode (no echo, no edit).  
CR  
A carriage return is required after each command line is typed into the terminal/computer. The  
command will not be sent to the analyzer to be executed until this is done. On personal  
computers, this is achieved by pressing the ENTER key.  
(carriage return)  
BS  
Erases one character to the left of the cursor location.  
(backspace)  
ESC  
Erases the entire command line.  
(escape)  
? [ID] CR  
This command prints a complete list of available commands along with the definitions of their  
functionality to the display device of the terminal or computer being used. The ID number of  
the analyzer is only necessary if multiple analyzers are on the same communications line, such  
as the multi-drop setup.  
Control-C  
Control-P  
Pauses the listing of commands.  
Restarts the listing of commands.  
126  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Model 360E Instruction Manual  
Operating Instructions  
6.15.2.3. Command Syntax  
Commands are not case-sensitive and all arguments within one command (i.e. ID numbers, keywords, data  
values, etc.) must be separated with a space character.  
All Commands follow the syntax:  
X [ID] COMMAND <CR>  
Where  
X
is the command type (one letter) that defines the type of command. Allowed designators  
are listed in Table 6-27 and Appendix A-6.  
[ID]  
is the analyzer identification number (Section 6.11.1.). Example: the Command “? 200”  
followed by a carriage return would print the list of available commands for the revision of  
software currently installed in the instrument assigned ID Number 200.  
COMMAND is the command designator: This string is the name of the command being issued (LIST,  
ABORT, NAME, EXIT, etc.). Some commands may have additional arguments that define  
how the command is to be executed. Press ? <CR> or refer to Appendix A-6 for a list of  
available command designators.  
<CR>  
is a carriage return. All commands must be terminated by a carriage return (usually  
achieved by pressing the ENTER key on a computer).  
Table 6-27: Command Types  
COMMAND  
COMMAND TYPE  
Calibration  
Diagnostic  
C
D
L
Logon  
T
Test measurement  
Variable  
V
W
Warning  
6.15.2.4. Data Types  
Data types consist of integers, hexadecimal integers, floating-point numbers, Boolean expressions and text  
strings.  
Integer data are used to indicate integral quantities such as a number of records, a filter length, etc.  
They consist of an optional plus or minus sign, followed by one or more digits. For example, +1, -12,  
123 are all valid integers.  
Hexadecimal integer data are used for the same purposes as integers. They consist of the two  
characters “0x,” followed by one or more hexadecimal digits (0-9, A-F, a-f), which is the ‘C’ programming  
language convention. No plus or minus sign is permitted. For example, 0x1, 0x12, 0x1234abcd are all  
valid hexadecimal integers.  
Floating point numbers are used to specify continuously variable values such as temperature set points,  
time intervals, warning limits, voltages, etc. They consist of an optional plus or minus sign, followed by  
zero or more digits, an optional decimal point, and zero or more digits. (At least one digit must appear  
before or after the decimal point.) Scientific notation is not permitted. For example, +1.0, 1234.5678, -  
0.1, 1 are all valid floating-point numbers.  
05232 Rev B3  
127  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Operating Instructions  
Model 360E Instruction Manual  
Boolean expressions are used to specify the value of variables or I/O signals that may assume only two  
values. They are denoted by the keywords ON and OFF.  
Text strings are used to represent data that cannot be easily represented by other data types, such as  
data channel names, which may contain letters and numbers. They consist of a quotation mark,  
followed by one or more printable characters, including spaces, letters, numbers, and symbols, and a  
final quotation mark. For example, “a”, “1”, “123abc”, and “()[]<>” are all valid text strings. It is not  
possible to include a quotation mark character within a text string.  
Some commands allow you to access variables, messages, and other items, such as iDAS data  
channels, by name. When using these commands, you must type the entire name of the item; you  
cannot abbreviate any names.  
6.15.2.5. Status Reporting  
Reporting of status messages as an audit trail is one of the three principal uses for the RS-232 interface (the  
other two being the command line interface for controlling the instrument and the download of data in electronic  
format). You can effectively disable the reporting feature by setting the interface to quiet mode (Section 6.11.5.,  
Table 6-10).  
Status reports include iDAS data (when reporting is enabled), warning messages, calibration and diagnostic  
status messages. Refer to Appendix A-3 for a list of the possible messages, and this for information on  
controlling the instrument through the RS-232 interface.  
General Message Format  
All messages from the instrument (including those in response to a command line request) are in the format:  
X DDD:HH:MM [Id] MESSAGE<CRLF>  
Where:  
X
is a command type designator, a single character indicating the message type, as  
shown in the Table 6-27.  
DDD:HH:MM is the time stamp, the date and time when the message was issued. It consists of the  
Day-of-year (DDD) as a number from 1 to 366, the hour of the day (HH) as a number  
from 00 to 23, and the minute (MM) as a number from 00 to 59.  
[ID]  
is the analyzer ID, a number with 1 to 4 digits.  
MESSAGE  
is the message content that may contain warning messages, test measurements, iDAS  
reports, variable values, etc.  
<CRLF>  
is a carriage return / line feed pair, which terminates the message.  
The uniform nature of the output messages makes it easy for a host computer to parse them into an easy  
structure. Keep in mind that the front panel display does not give any information on the time a message was  
issued, hence it is useful to log such messages for trouble-shooting and reference purposes. Terminal  
emulation programs such as HyperTerminal can capture these messages to text files for later review.  
128  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 360E Instruction Manual  
Operating Instructions  
6.15.2.6. Remote Access by Modem  
The M360E can be connected to a modem for remote access. This requires a cable between the analyzer’s  
COM port and the modem, typically a DB-9F to DB-25M cable (available from Teledyne Instruments with part  
number WR0000024).  
Once the cable has been connected, check to make sure the DTE-DCE is in the correct position. Also make  
sure the M360E COM port is set for a baud rate that is compatible with the modem, which needs to operate with  
an 8-bit word length with one stop bit.  
The first step is to turn on the MODEM ENABLE communication mode (Mode 64, Section 6.11.5). Once this is  
completed, the appropriate setup command line for your modem can be entered into the analyzer. The default  
setting for this feature is  
AT Y0 &D0 &H0 &I0 S0=2 &B0 &N6 &M0 E0 Q1 &W0  
This string can be altered to match your modem’s initialization and can be up to 100 characters long.  
To change this setting press:  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
SETUP X.X  
SET> EDIT  
COM1 MODE:0  
EXIT  
EXIT  
SETUP X.X  
PRIMARY SETUP MENU  
CFG DAS RNGE PASS CLK MORE  
EXIT  
EXIT  
SETUP X.X  
COM1 BAUD RATE:19200  
<SET SET> EDIT  
EXIT returns to  
the previous  
menu  
SETUP X.X SECONDARY SETUP MENU  
COMM VARS DIAG ALRM  
SETUP X.X  
COM1 MODEM INIT:AT Y&D&H  
EXIT  
<SET SET> EDIT  
SETUP X.X  
COMMUNICATIONS MENU  
Select which  
COM Port is  
tested  
ENTR accepts the  
new string and returns  
to the previous menu.  
EXIT ignores the new  
string and returns to  
the previous menu.  
ID COM1 COM2  
EXIT  
SETUP X.X  
COM1 MODEM INIT:[A]T Y&D&H  
ENTR EXIT  
<CH CH> INS DEL [A]  
Press the [?]  
key repeatedly to cycle through the  
available character set:  
0-9  
The INS key  
inserts a character  
before the cursor  
location.  
The DEL key  
deletes a character  
at the cursor  
location.  
A-Z  
The <CH and CH> keys move  
the [ ] cursor left and right  
along the text string  
space ’ ~ ! # $ % ^ & * ( ) - _ =  
+[ ] { } < >\ | ; : , . / ?  
05232 Rev B3  
129  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Operating Instructions  
Model 360E Instruction Manual  
To initialize the modem press:  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
SETUP X.X  
SET> EDIT  
COM1 MODE:0  
EXIT  
SETUP X.X  
PRIMARY SETUP MENU  
CFG DAS RNGE PASS CLK MORE  
EXIT  
SETUP X.X  
COM1 BAUD RATE:19200  
EXIT returns to  
the previous  
menu  
<SET SET> EDIT  
EXIT  
SETUP X.X SECONDARY SETUP MENU  
COMM VARS DIAG ALRM  
EXIT  
SETUP X.X  
COM1 MODEM INIT:AT Y&D&H  
SETUP X.X  
COMMUNICATIONS MENU  
<SET SET> EDIT  
EXIT  
Select which  
COM Port is  
tested  
ID COM1 COM2  
EXIT  
SETUP X.X  
COM1 INITIALIZE MODEM  
EXIT  
<SET SET> INIT  
SETUP X.X  
INITIALIZING MODEM  
<SET SET> INIT  
EXIT  
EXIT returns to the  
Communications Menu.  
6.15.2.7. COM Port Password Security  
In order to provide security for remote access of the M360E, a LOGON feature can be enabled to require a  
password before the instrument will accept commands. This is done by turning on the SECURITY MODE (Mode  
4, Section 6.11.5). Once the SECURITY MODE is enabled, the following items apply.  
A password is required before the port will respond or pass on commands.  
If the port is inactive for one hour, it will automatically logoff, which can also be achieved with the  
LOGOFF command.  
Three unsuccessful attempts to log on with an incorrect password will cause subsequent logins to be  
disabled for 1 hour, even if the correct password is used.  
If not logged on, the only active command is the '?' request for the help screen.  
The following messages will be returned at logon:  
LOGON SUCCESSFUL - Correct password given  
LOGON FAILED - Password not given or incorrect  
LOGOFF SUCCESSFUL - Connection terminated successfully  
To log on to the model 360E analyzer with SECURITY MODE feature enabled, type:  
LOGON 940331  
940331 is the default password. To change the default password, use the variable RS232_PASS issued as  
follows:  
V RS232_PASS=NNNNNN  
Where N is any numeral between 0 and 9.  
130  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 360E Instruction Manual  
Operating Instructions  
6.15.2.8. APICOM Remote Control Program  
APICOM is an easy-to-use, yet powerful interface program that allows accessing and controlling any of Teledyne  
Instruments’ main line of ambient and stack-gas instruments from a remote connection through direct cable,  
modem or Ethernet. Running APICOM, a user can:  
Establish a link from a remote location to the M360E through direct cable connection via RS-232 modem or  
Ethernet.  
View the instrument’s front panel and remotely access all functions that could be accessed when  
standing in front of the instrument.  
Remotely edit system parameters and set points.  
Download, view, graph and save data for predictive diagnostics or data analysis.  
Retrieve, view, edit, save and upload iDAS configurations.  
Check on system parameters for trouble-shooting and quality control.  
APICOM is very helpful for initial setup, data analysis, maintenance and trouble-shooting. Figure 6-5 shows an  
example of APICOM being used to remotely configuration the instruments iDAS feature. Figure 6-19 shows  
examples of APICOM’s main interface, which emulates the look and functionality of the instruments actual front  
panel  
Figure 6-19: APICOM Remote Control Program Interface  
APICOM is included free of cost with the analyzer and the latest versions can also be downloaded for free at  
http://www.teledyne-api.com/software/apicom/.  
6.15.3. Additional Communications Documentation  
Table 6-28: Serial Interface Documents  
INTERFACE / TOOL  
DOCUMENT TITLE  
PART  
AVAILABLE ONLINE*  
NUMBER  
APICOM  
Multi-drop  
DAS Manual  
APICOM User Manual  
039450000  
021790000  
028370000  
YES  
YES  
YES  
RS-232 Multi-drop Documentation  
Detailed description of the iDAS.  
* These documents can be downloaded at http://www.teledyne-api.com/manuals/  
05232 Rev B3  
131  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Operating Instructions  
Model 360E Instruction Manual  
6.15.4. Using the M360E with a Hessen Protocol Network  
6.15.4.1. General Overview of Hessen Protocol  
The Hessen protocol is a multidrop protocol, in which several remote instruments are connected via a common  
communications channel to a host computer. The remote instruments are regarded as slaves of the host  
computer. The remote instruments are unaware that they are connected to a multidrop bus and never initiate  
Hessen protocol messages. They only respond to commands from the host computer and only when they  
receive a command containing their own unique ID number.  
The Hessen protocol is designed to accomplish two things: to obtain the status of remote instruments, including  
the concentrations of all the gases measured; and to place remote instruments into zero or span calibration or  
measure mode. API’s implementation supports both of these principal features.  
The Hessen protocol is not well defined, therefore while API’s application is completely compatible with the  
protocol itself, it may be different from implementations by other companies.  
The following subs describe the basics for setting up your instrument to operate over a Hessen Protocol  
network. For more detailed information as well as a list of host computer commands and examples of command  
and response message syntax, download the Manual Addendum for Hessen Protocol from the Teledyne  
Instruments web site: http://www.teledyne-api.com/manuals/index.asp .  
6.15.4.2. Hessen COMM Port Configuration  
Hessen protocol requires the communication parameters of the M360E’s COMM ports to be set differently than  
the standard configuration as shown in the table below.  
Table 6-29: RS-232 Communication Parameters for Hessen Protocol  
Parameter  
Data Bits  
Stop Bits  
Parity  
Standard  
Hessen  
7
8
1
2
None  
Full  
Even  
Half  
Duplex  
To change the rest of the COMM port parameters. See Section 6.11.5.  
To change the baud rate of the M360E’s COMM ports, See Section 6.11.8.  
NOTE  
Make sure that the communication parameters of the host computer are also properly  
set.  
Also, the instrument software has a 200 ms. Latency before it responds to commands  
issued by the host computer. This latency should present no problems, but you should  
be aware of it and not issue commands to the instrument too frequently.  
132  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Model 360E Instruction Manual  
Operating Instructions  
6.15.4.3. Activating Hessen Protocol  
The first step in configuring the M360E to operate over a Hessen protocol network is to activate the Hessen  
mode for COMM ports and configure the communication parameters for the port(s) appropriately. Press:  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
SETUP X.X  
COM1 QUIET MODE: OFF  
ENTR EXIT  
Repeat the entire  
process to set up  
the COM2 port  
< TST TST > CAL  
SETUP  
NEXT OFF  
SETUP X.X  
PRIMARY SETUP MENU  
Continue pressing next until …  
CFG DAS RNGE PASS CLK MORE  
EXIT  
EXIT  
EXIT  
EXIT  
SETUP X.X COM1 HESSEN PROTOCOL : OFF  
SETUP X.X SECONDARY SETUP MENU  
PREV NEXT OFF  
ENTR EXIT  
COMM VARS DIAG  
ALRM  
Toggle OFF/ON keys to  
change  
activate/deactivate  
SETUP X.X COM1 HESSEN PROTOCOL : ON  
SETUP X.X  
COMMUNICATIONS MENU  
Select which COMM  
port to configure  
selected mode.  
PREV NEXT ON  
ENTR EXIT  
ID COM1 COM2  
The sum of the mode  
IDs of the selected  
SETUP X.X  
COM1 E,7,1 MODE: OFF  
SETUP X.X  
SET> EDIT  
COM1 MODE:0  
modes is displayed here  
PREV NEXT OFF  
ENTR EXIT  
SETUP X.X  
COM1 E,7,1 MODE: ON  
ENTR key accepts the  
new settings  
EXIT key ignores the new  
PREV NEXT ON  
ENTR EXIT  
settings  
6.15.4.4. Selecting a Hessen Protocol Type  
Currently there are two versions of Hessen Protocol in use. The original implementation, referred to as TYPE 1,  
and a more recently released version, TYPE 2 that has more flexibility when operating with instruments that can  
measure more than one type of gas. For more specific information about the difference between TYPE 1and  
TYPE 2 download the Manual Addendum for Hessen Protocol from the Teledyne Instruments web site:  
http://www.teledyne-api.com/manuals/index.asp .  
To select a Hessen Protocol Type press:  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
SETUP X.X  
COMMUNICATIONS MENU  
SETUP X.X  
PRIMARY SETUP MENU  
ID HESN COM1 COM2  
CFG DAS RNGE PASS CLK MORE  
EXIT  
EXIT  
SETUP X.  
HESSEN VARIATION: TYPE 1  
SETUP X.X SECONDARY SETUP MENU  
SET> EDIT  
EXIT  
COMM VARS DIAG ALRM  
ENTR key accepts the  
new settings  
SETUP X.X HESSEN VARIATION: TYPE 1  
TYE1 TYPE 2 ENTR EXIT  
EXIT key ignores the new  
settings  
SETUP X.X HESSEN VARIATION: TYPE 2  
PREV NEXT OFF ENTR EXIT  
Press to change  
protocol type.  
05232 Rev B3  
133  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Operating Instructions  
Model 360E Instruction Manual  
NOTE  
While Hessen Protocol Mode can be activated independently for COM1 and COM2, The  
TYPE selection affects both Ports.  
6.15.4.5. Setting The Hessen Protocol Response Mode  
The Teledyne Instruments implementation of Hessen Protocol allows the user to choose one of several different  
modes of response for the analyzer.  
Table 6-30: Teledyne Instruments Hessen Protocol Response Modes  
MODE ID  
CMD  
MODE DESCRIPTION  
This is the Default Setting. Reponses from the instrument are encoded as the traditional command format.  
Style and format of responses depend on exact coding of the initiating command.  
Responses from the instrument are always delimited with <STX> (at the beginning of the response, <ETX>  
(at the end of the response followed by a 2 digit Block Check Code (checksum), regardless of the command  
encoding.  
BCC  
Responses from the instrument are always delimited with <CR> at the beginning and the end of the string,  
regardless of the command encoding.  
TEXT  
To Select a Hessen response mode, press:  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
SETUP X.X  
COMMUNICATIONS MENU  
ID HESN COM1 COM2  
EXIT  
EXIT  
SETUP X.X  
PRIMARY SETUP MENU  
CFG DAS RNGE PASS CLK MORE  
EXIT  
EXIT  
SETUP X.X  
HESSEN VARIATION: TYPE 1  
SET> EDIT  
SETUP X.X SECONDARY SETUP MENU  
ENTR key accepts the  
new settings  
COMM VARS DIAG ALRM  
EXIT key ignores the new  
SETUP X.X  
HESSEN RESPONSE MODE :CMD  
settings  
<SET SET> EDIT  
EXIT  
Press to  
change  
response  
mode.  
SETUP X.X  
HESSEN RESPONSE MODE :CMD  
BCC TEXT EDIT  
ENTR EXIT  
6.15.4.6. Hessen Protocol Gas ID  
The Model 360E Analyzer is a single gas instrument that measures CO2. As such it’s default gas ID has already  
been set to 310. There is no need to change this setting.  
134  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Model 360E Instruction Manual  
Operating Instructions  
6.15.4.7. Setting Hessen Protocol Status Flags  
Teledyne Instruments’ implementation of Hessen protocols includes a set of status bits that the instrument  
includes in responses to inform the host computer of its condition. Each bit can be assigned to one operational  
and warning message flag. The default settings for these bit/flags are:  
Table 6-31: Default Hessen Status Bit Assignments  
STATUS FLAG NAME  
DEFAULT BIT  
ASSIGNMENT  
WARNING FLAGS  
SAMPLE FLOW WARNING  
BENCH TEMP WARNING  
SOURCE WARNING  
0001  
0002  
0004  
0008  
0010  
0020  
0040  
0080  
BOX TEMP WARNING  
WHEEL TEMP WARNING  
SAMPLE TEMP WARNING  
SAMPLE PRESSURE WARNING  
INVALID CONC  
(The Instrument’s Front Panel Display Will Show The  
Concentration As “XXXX”)  
OPERATIONAL FLAGS  
Instrument Off  
0100  
0200  
0400  
0800  
In Manual Calibration Mode  
In Zero Calibration Mode  
In Span Calibration Mode  
UNITS OF MEASURE FLAGS  
UGM  
0000  
2000  
MGM  
PPB  
4000  
PPM  
6000  
SPARE/UNUSED BITS  
UNASSIGNED FLAGS (0000)  
100, 1000, 8000  
Sync Warning  
Relay Board Warning  
Conc Alarm 1  
Front Panel Warning  
Analog Cal Warning  
Cannot Dyn Zero  
Cannot Dyn Span  
Invalid Conc  
Conc Alarm 2  
Photo Temp Warning  
System Reset  
Rear Board Not Detected  
NOTES:  
It is possible to assign more than one flag to the same Hessen status bit. This allows  
the grouping of similar flags, such as all temperature warnings, under the same status  
bit.  
Be careful not to assign conflicting flags to the same bit as each status bit will be  
triggered if any of the assigned flags is active.  
05232 Rev B3  
135  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Operating Instructions  
Model 360E Instruction Manual  
To assign or reset the status flag bit assignments, press:  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
SETUP X.X  
PRIMARY SETUP MENU  
CFG DAS RNGE PASS CLK MORE  
EXIT  
EXIT  
SETUP X.X SECONDARY SETUP MENU  
COMM VARS DIAG ALRM  
SETUP X.X  
COMMUNICATIONS MENU  
ID HESN COM1 COM2  
EXIT  
Repeat pressing SET> until …  
SETUP X.  
HESSEN STATUS FLAGS  
<SET SET> EDIT  
EXIT  
SETUP X.  
SYNC WARNING: 0000  
PREV NEXT  
EDIT PRNT EXIT  
Repeat pressing NEXT or PREV until the desired  
message flag is displayed. See Table 6-27.  
For xxample …  
SETUP X.  
SYSTEM RESET: 0000  
EDIT PRNT EXIT  
PREV NEXT  
The <CH and  
CH> keys move  
the [ ] cursor left  
and right along  
the bit string.  
SETUP X.  
SYSTEM RESET: [0]000  
[0]  
ENTR key accepts the  
new settings  
<CH CH>  
ENTR EXIT  
EXIT key ignores the new  
settings  
Press the [?] key repeatedly to cycle through the available character set: 0-9  
Note: Values of A-F can also be set but are meaningless.  
6.15.4.8. Instrument ID Code  
Each instrument on a Hessen Protocol network must have a unique ID code. The M360E is programmed with a  
default ID code of 360. To change this code See Section 6.11.1  
User Notes  
136  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 360E Instruction Manual  
Calibration Procedures  
7. CALIBRATION PROCEDURES  
This contains a variety of information regarding the various methods for calibrating a Model 360E CO2 Analyzer  
as well as other supporting information.  
NOTE  
The procedures in this assume that the calibration password feature is disabled (the  
instruments default state). If it is enabled a password prompt screen (See Section 6.9)  
will appear after the CAL, CALZ or CALS buttons are pushed but before the instrument  
enters the associated calibration mode.  
7.1. Before Calibration  
The calibration procedures in this assume that the Range Type, Range Span and units of measure have already  
been selected for the analyzer. If this has not been done, please do so before continuing (See Section 6.8 for  
instructions).  
All Gas lines should be PTFE (Teflon), FEP, glass, stainless steel or brass.  
NOTE  
If any problems occur while performing the following calibration procedures, refer to  
Chapter 11 of this manual for troubleshooting tips.  
7.1.1. Zero Air and Span Gas  
To perform the following calibration you must have sources for zero air and span gas available.  
Zero Air is similar in chemical composition to the Earth’s atmosphere but scrubbed of all components that might  
affect the analyzer’s readings. Zero air should contain less than 25 ppb of CO2 and other major interfering  
gases such as CO and Water Vapor. It should have a dew point of -5C or less  
Span Gas is a gas specifically mixed to match the chemical composition of the type of gas being measured at  
near full scale of the desired measurement range. It is recommended that the span gas used have a  
concentration equal to 80% of the full measurement range.  
If Span Gas is sourced directly from a calibrated, pressurized tank, the gas mixture should be CO2 mixed with  
Zero Air or N2 at the required ratio.  
Zero air generators that condition ambient air by drying and removal of pollutants are available on the  
commercial market such as the Teledyne Instruments Model 701 Zero Air Generator. We recommend this type  
of device, in conjunction with a CO2 scrubber such as soda lime (such as Teledyne Instruments P/N  
037600000), for generating zero air.  
05232 Rev B3  
137  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Calibration Procedures  
Model 360E Instruction Manual  
7.1.2. Calibration Gas Traceability  
All equipment used to produce calibration gases should be verified against standards of the National Institute for  
Standards and Technology (NIST). To ensure NIST traceability, we recommend acquiring cylinders of working  
gas that are certified to be traceable to NIST Standard Reference Materials (SRM). These are available from a  
variety of commercial sources.  
7.1.3. Data Recording Devices  
A strip chart recorder, data acquisition system or digital data acquisition system should be used to record data  
from the M360E’s serial or analog outputs. If analog readings are used, the response of the recording system  
should be checked against a NIST traceable voltage source or meter. Data recording device should be capable  
of bi-polar operation so that negative readings can be recorded. For electronic data recording, the M360E  
provides an internal data acquisition system (iDAS), which is described in detail in Section 6.7.  
7.2. Manual Calibration without Zero/Span Valves  
This is the basic method for manually calibrating the Model 360E CO2 Analyzer without functioning zero/span  
valve options. It is identical to the method described in the GETTING STARTED (Chapter 3) of this manual and  
is repeated her for you convenience.  
STEP ONE: Connect the Sources of Zero Air and Span Gas as shown below.  
VENT here if input  
is pressurized  
Source of  
SAMPLE GAS  
Removed during  
calibration  
Calibrated  
CO2 Gas  
at span gas  
concentration  
Indicating  
SAMPLE  
Soda Lime  
VENT  
EXHAUST  
VENT SPAN  
PRESSURE SPAN  
IZS  
MODEL  
360E  
MODEL 701  
Zero Gas  
PURGE LINE  
Generator  
Figure 7-1: Pneumatic Connections–Basic Configuration–Using Bottled Span Gas  
138  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Model 360E Instruction Manual  
Calibration Procedures  
Figure 7-2: Pneumatic Connections–Basic Configuration–Using Gas Dilution Calibrator  
STEP TWO: Set the expected CO2 Span Gas concentration:  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
SETUP  
< TST TST > CAL  
This sequence causes the  
analyzer to prompt for the  
expected CO2 span  
concentration.  
M-P CAL  
RANGE = 500.000 PPM  
CO2 =X.XXX  
EXIT  
< TST TST > ZERO  
CONC  
The CO2 span  
concentration values  
automatically default to  
400.0 Conc.  
EXIT ignores the new setting  
and returns to the previous  
display.  
M-P CAL  
CO2 SPAN CONC: 400.000 Conc  
To change this value to  
the actual concentration of  
the span gas, enter the  
number by pressing the  
key under each digit until  
the expected value  
ENTR accepts the new setting  
0
0
0
4
5
.0  
ENTR EXIT  
and returns to the  
previous display..  
appears.  
NOTE  
For this Initial Calibration it is important to independently verify the PRECISE CO2  
Concentration Value of the SPAN gas.  
If the source of the Span Gas is from a Calibrated Bottle, use the exact concentration  
value printed on the bottle.  
05232 Rev B3  
139  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Calibration Procedures  
Model 360E Instruction Manual  
STEP THREE: Perform the Zero/Span Calibration Procedure:  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =XXX.X  
SETUP  
Set the Display to show the  
STABIL test function.  
This function calculates the  
stability of the CO2  
< TST TST > CAL  
measurement  
SAMPLE  
STABIL=XXX.X PPM  
CO2 =XXX.X  
SETUP  
< TST TST > CAL  
ACTION:  
Allow zero gas to enter the sample port at the  
rear of the instrument.  
Wait until STABIL  
falls below 1.0 ppm.  
This may take several  
minutes.  
M-P CAL  
STABIL=XXX.X PPM  
CO2 =XXX.X  
< TST TST > CAL  
SETUP  
M-P CAL  
STABIL=XXX.X PPM  
CONC  
CO2 =XXX.X  
EXIT  
< TST TST > ZERO  
Press ENTR to changes the  
OFFSET & SLOPE values for the  
CO2 measurements.  
M-P CAL  
STABIL=XXX.X PPM  
CO2 =XXX.X  
< TST TST > ENTR  
CONC  
EXIT  
Press EXIT to leave the calibration  
unchanged and return to the  
previous menu.  
ACTION:  
Allow span gas to enter the sample port at the  
rear of the instrument.  
The value of  
STABIL may jump  
significantly.  
Wait until it falls back  
below 1.0 ppm  
The SPAN key now  
appears during the  
transition from zero to  
span.  
This may take several  
minutes.  
M-P CAL  
STABIL=XXX.X PPM  
CO2 =XXX.X  
EXIT  
< TST TST >  
M-P CAL  
SPAN CONC  
You may see both keys.  
If either the ZERO or  
SPAN buttons fail to  
appear see Section 11  
for troubleshooting tips.  
Press ENTR to change the  
OFFSET & SLOPE values for the  
CO2 measurements.  
RANGE = 500.000 PPM CO2 =XXX.X  
EXIT  
Press EXIT to leave the calibration  
unchanged and return to the  
previous menu.  
< TST TST > ENTR SPAN CONC  
M-P CAL  
RANGE = 500.000 PPM CO2 =XXX.X  
CONC EXIT  
EXIT returns to the main  
SAMPLE display  
< TST TST > ENTR  
If the ZERO or SPAN keys are not displayed, this means that the measurement made during that part of the  
procedure is too far out of the allowable range to do allow a reliable calibration. The reason for this must be  
determined before the analyzer can be calibrated. See Chapter 11 for troubleshooting tips.  
140  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual  
Calibration Procedures  
7.3. Manual Calibration Checks  
Informal calibration checks, which only evaluate but do not alter the analyzer’s response curve, are  
recommended as a regular maintenance item and in order to monitor the analyzer’s performance. To carry out a  
calibration check rather than a full calibration, follow these steps.  
STEP ONE: Connect the sources of zero air and span gas as shown in Figures 7.1 or 7.2.  
STEP TWO: Perform the zero/span calibration check procedure:  
ACTION:  
Supply the instrument with zero gas.  
SAMPLE  
RANGE = 500.0 PPM  
CO2=X.XXX  
SETUP  
Scroll the display to the  
STABIL test function.  
< TST TST > CAL  
SAMPLE  
STABIL=XXX.X PPM  
CO2=X.XXX  
SETUP  
< TST TST > CAL  
Wait until  
STABIL is  
below 1.0 ppm.  
This may take  
several minutes.  
ACTION:  
Record the CO2  
concentration  
reading.  
SAMPLE  
STABIL=XXX.X PPM  
CO2=X.XXX  
< TST TST > CAL  
SETUP  
The value of  
STABIL may jump  
significantly.  
ACTION:  
Supply span gas to the instrument  
Wait until it falls  
below 1.0 ppm. This  
may take several  
minutes.  
ACTION:  
Record the CO2  
concentration  
reading.  
SAMPLE  
STABIL=XXX.X PPM  
CO2=X.XXX  
SETUP  
< TST TST > CAL  
The SPAN key appears during the transition from zero to  
span. You may see both keys.  
7.4. Manual Calibration with Zero/Span Valves  
There are four different zero/span valve option configurations (See Section 5.4). They all operate identically,  
differing only in the method used to supply calibration gas to the Analyzer.  
STEP ONE: Connect the sources of Zero Air and Span Gas as shown below.  
Figures 7-3 through 7-6 show the proper pneumatic connections for M360E’s with various optional internal valve  
sets installed.  
05232 Rev B3  
141  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Calibration Procedures  
Model 360E Instruction Manual  
Figure 7-3: Pneumatic Connections–M360E with Zero/Span/Shutoff Valves (OPT 50)  
VENT here if input  
is pressurized  
Source of  
SAMPLE GAS  
Removed during  
calibration  
SAMPLE  
EXHAUST  
VENT SPAN  
VENT  
MODEL  
360E  
Calibrated  
CO2 Gas  
at span gas  
concentration  
PRESSURE SPAN  
External  
Zero Air  
Scrubber  
IZS  
PURGE LINE  
MODEL 701  
Zero Gas  
Indicating  
Soda Lime  
Generator  
Figure 7-4  
Pneumatic Connections–M360E with Zero/Span/Shutoff Valves and External Zero Air  
Scrubber (OPT 51)  
142  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Model 360E Instruction Manual  
Calibration Procedures  
Figure 7-5: Pneumatic Connections–M360E with Zero/Span Valves (OPT 52)  
Figure 7-6: Pneumatic Connections–M360E with Zero/Span Valves with External Zero air Scrubber  
(OPT 53)  
05232 Rev B3  
143  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Calibration Procedures  
Model 360E Instruction Manual  
STEP TWO: Set the expected CO2 Span Gas concentration:  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
SETUP  
< TST TST > CAL CALZ CALS  
This sequence causes the  
analyzer to prompt for the  
expected CO2 span  
concentration.  
M-P CAL  
RANGE = 500.000 PPM  
CONC  
CO2 =X.XXX  
EXIT  
< TST TST > ZERO  
The CO2 span  
concentration values  
automatically default to  
400.0 Conc.  
EXIT ignores the new setting  
and returns to the previous  
display.  
M-P CAL  
CO2 SPAN CONC: 450.000 Conc  
To change this value to  
the actual concentration of  
the span gas, enter the  
number by pressing the  
key under each digit until  
the expected value  
ENTR accepts the new setting  
0
0
0
4
5
.0  
ENTR EXIT  
and returns to the  
previous display.  
appears.  
NOTE  
For this Initial Calibration it is important to independently verify the PRECISE CO2  
Concentration Value of the SPAN gas.  
If the source of the Span Gas is from a Calibrated Bottle, use the exact concentration  
value printed on the bottle.  
144  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual  
Calibration Procedures  
STEP THREE: Perform the zero/span calibration. Zero and span checks using the zero/span valve option are  
similar to that described in Section 7.2, except that zero air and span gas is supplied to the analyzer through the  
zero/span valves rather than through the sample inlet port.  
The zero and cal operations are initiated directly and independently with dedicated keys (CALZ & CALS).  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =XXX.X  
Set the Display to show the  
STABIL test function.  
This function calculates the  
stability of the COx  
<TST TST> CAL CALZ CALS SETUP  
measurement  
SAMPLE  
STABIL=XXX.X PPM  
CO2 =XXX.X  
< TST TST > CAL CALZ CALS SETUP  
ACTION:  
Allow zero gas to enter the sample port at the  
rear of the instrument.  
Wait until STABIL  
falls below1.0 ppm.  
This may take several  
minutes.  
M-P CAL  
STABIL=XXX.X PPM  
CO2 =XXX.X  
SETUP  
< TST TST > CAL CALZ CALS  
M-P CAL  
STABIL=XXX.X PPM  
CONC  
CO2 =XXX.X  
EXIT  
< TST TST > ZERO  
Press ENTR to changes the  
OFFSET & SLOPE values for the  
CO2 measurements.  
M-P CAL  
STABIL=XXX.X PPM  
CO2 =XXX.X  
< TST TST > ENTR  
CONC  
EXIT  
Press EXIT to leave the calibration  
unchanged and return to the  
previous menu.  
ACTION:  
Allow span gas to enter the sample port at the  
rear of the instrument.  
The value of  
STABIL may jump  
significantly.  
Wait until it falls back  
below 1.0 ppm.  
The SPAN key now  
appears during the  
transition from zero to  
span.  
This may take several  
minutes.  
M-P CAL  
STABIL=XXX.X PPM  
CO2 =XXX.X  
EXIT  
< TST TST >  
M-P CAL  
SPAN CONC  
You may see both keys.  
If either the ZERO or  
SPAN buttons fail to  
appear see Section 11  
for troubleshooting tips.  
Press ENTR to change the  
OFFSET & SLOPE values for the  
CO2 measurements.  
RANGE = 500.000 PPM CO2 =XXX.X  
EXIT  
Press EXIT to leave the calibration  
unchanged and return to the  
previous menu.  
< TST TST > ENTR SPAN CONC  
M-P CAL  
RANGE = 500.000 PPM CO2 =XXX.X  
CONC EXIT  
EXIT returns to the main  
SAMPLE display  
< TST TST > ENTR  
05232 Rev B3  
145  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Calibration Procedures  
Model 360E Instruction Manual  
7.5. Manual Calibration Checks with Zero/Span Valves  
Zero and span checks using the VARIOUS zero/span valve options available for the M360E are similar to that  
described in Section 7.3, except that the zero and calibration operations are initiated directly and independently  
with dedicated keys CALZ and CALS.  
To perform a manual calibration check of an analyzer with a valve option installed, use the following method.  
STEP ONE: Connect the sources of Zero Air and Span Gas as shown in Figures 7-3 through 7-6.  
STEP TWO: Perform the zero/span check.  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
Scroll to the STABIL test  
< TST TST > CAL CALZ CALS  
SETUP  
function.  
SAMPLE  
STABIL=XXX.X PPM  
CO2 =X.XXX  
SETUP  
Wait until STABIL  
falls below 1.0  
ppm.  
< TST TST > CAL CALZ CALS  
ACTION:  
Record the  
CO2 readings  
presented in the  
upper right corner of  
the display.  
This may take  
several minutes.  
ZERO CAL M  
STABIL=XXX.X PPM CO2 =X.XXX  
CONC EXIT  
< TST TST > ZERO  
SAMPLE  
STABIL=XXX.X PPM CO2 =X.XXX  
ACTION:  
Record the  
CO2 readings  
presented in the  
upper right corner of  
the display.  
The value of STABIL  
may jump  
significantly. Wait  
until STABIL falls  
below 1.0 ppm. This  
may take several  
minutes.  
< TST TST > CAL CALZ CALS  
SETUP  
SPAN CAL M  
STABIL=XXX.X PPM  
CO2 =X.XXX  
EXIT  
EXIT returns to the main  
< TST TST > ZERO SPAN CONC  
SAMPLE display  
146  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 360E Instruction Manual  
Calibration Procedures  
7.5.1. Zero/Span Calibration on Auto Range or Dual Ranges  
If the analyzer is being operated in dual range mode or auto range mode, then the high and low ranges must be  
independently calibrated.  
When the analyzer is in either dual or auto range modes the user must run a separate calibration procedure for  
each range. After pressing the CAL, CALZ or CALS keys the user is prompted for the range that is to be  
calibrated as seen in the CALZ example below:  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =XXX.X  
SETUP  
Set the Display to show the  
STABIL test function.  
This function calculates the  
stability of the CO2  
<TST TST> CAL CALZ CALS  
measurement  
SAMPLE  
STABIL=XXX.X PPM  
CO2 =XXX.X  
SETUP  
< TST TST > CAL CALZ CALS  
ACTION:  
Allow zero gas to enter the sample port at the  
rear of the instrument.  
Wait until STABIL  
falls below 1.0 ppm.  
This may take several  
minutes.  
M-P CAL  
STABIL=XXX.X PPM  
CO2 =XXX.X  
< TST TST > CAL CALZ CALS  
SETUP  
SAMPLE  
RANGE TO CAL: LOW  
ENTR  
LOW HIGH  
SETUP  
SETUP  
SAMPLE  
RANGE TO CAL: HIGH  
ENTR  
LOW HIGH  
ANALYZER ENTERS  
ZERO CAL MODE  
ZERO CAL M  
RANGE = 500.000 PPM CO2 =XXX.X  
< TST TST > ZERO SPAN CONC  
EXIT  
Continue Calibration as per  
Standard Procedure  
Once this selection is made, the calibration procedure continues as previously described in Section 7.2. The  
other range may be calibrated by starting over from the main SAMPLE display.  
05232 Rev B3  
147  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Calibration Procedures  
Model 360E Instruction Manual  
7.5.2. Use of Zero/Span Valves with Remote Contact Closure  
Contact closures for controlling calibration are located on the rear panel CONTROL IN connector. Instructions  
for setup and use of these contacts are found in Section 6.15.1.2. When the contacts are closed for at least 5  
seconds, the instrument switches into zero or span mode. The remote calibration contact closures may be  
activated in any order. It is recommended that contact closures remain closed for at least 10 minutes to  
establish a reliable reading.  
The instrument will stay in the selected mode for as long as the contacts remain closed. If calibration is enabled,  
the M360E will re-calibrate when the contact is opened, then go into SAMPLE mode. If calibration is disabled,  
the instrument will return to SAMPLE mode, leaving the calibration unchanged.  
7.6. Automatic Zero/Span Cal/Check (AutoCal)  
The AutoCal system allows unattended periodic operation of the ZERO/SPAN valve options by using the  
M360E’s internal time of day clock. AutoCal operates by executing SEQUENCES programmed by the user to  
initiate the various calibration modes of the analyzer and open and close valves appropriately. It is possible to  
program and run up to 3 separate sequences (SEQ1, SEQ2 and SEQ3). Each sequence can operate in one of  
3 Modes, or be disabled.  
Table 7-1: AUTOCAL Modes  
MODE NAME  
DISABLED  
ZERO  
ACTION  
Disables the Sequence  
Causes the Sequence to perform a zero calibration/check  
ZERO-SPAN  
Causes the Sequence to perform a zero and span concentration  
calibration/check  
SPAN  
Causes the Sequence to perform a span concentration  
calibration/check  
For each mode there are seven parameters that control operational details of the SEQUENCE. They are:  
Table 7-2: AutoCal ATTRIBUTE Setup Parameters  
ATTRIBUTE NAME  
TIMER ENABLED  
STARTING DATE  
STARTING TIME  
DELTA DAYS  
ACTION  
Turns on the Sequence timer  
Sequence will operate after Starting Date  
Time of day sequence will run  
Number of days to skip between each Seq. execution  
Number of hours later each “Delta Days” Seq is to be run  
Number of minutes the sequence operates  
DELTA TIME  
DURATION  
CALIBRATE  
Enable to do a calibration – Disable to do a cal check only  
148  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Model 360E Instruction Manual  
Calibration Procedures  
The following example sets Sequence #2 to do a Zero-Span Calibration every other day starting at 1 am on  
September 4, 2001, lasting 15 minutes, without calibration. This will start ½ hour later each iteration.  
MODE AND ATTRIBUTE  
Sequence  
VALUE  
COMMENT  
Define Sequence #2  
2
ZERO-SPAN  
ON  
Mode  
Select Zero and Span Mode  
Enable the timer  
Timer Enable  
Starting Date  
Starting Time  
Delta Days  
Delta Time  
Sept. 4, 2001  
01:00  
Start after Sept 4, 2001  
First Span starts at 1:00AM  
Do Sequence #2 every other day  
Do Sequence #2 ½ hr later each day  
Operate Span valve for 15 min  
Do not calibrate at end of Sequence  
2
00:30  
Duration  
15.0  
Calibrate  
NO  
NOTE  
The programmed STARTING_TIME must be a minimum of 5 minutes later than the real  
time clock (See Section 6.10 for setting real time clock).  
NOTE  
Avoid setting two or more sequences at the same time of the day. Any new sequence  
which is initiated whether from a timer, the COM ports, or the contact closure inputs  
will override any sequence which is in progress.  
NOTE  
If at any time an illegal entry is selected (Example: Delta Days > 367) the ENTR key will  
disappear from the display.  
05232 Rev B3  
149  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Calibration Procedures  
Model 360E Instruction Manual  
To program the Sequence:  
SAMPLE*  
RANGE = 500.000 PPM CO2 =X.XXX  
SETUP X.X STARTING TIME:14:15  
<SET SET> EDIT  
< TST TST > CAL CALZ CALS  
SETUP  
EXIT  
EXIT  
SETUP X.X  
PRIMARY SETUP MENU  
SETUP X.X  
DELTA DAYS: 1  
CFG ACAL DAS RNGE PASS CLK MORE  
EXIT  
EXIT  
<SET SET> EDIT  
Toggle keys  
to set  
number of  
days  
SETUP X.X SEQ 1) DISABLED  
SETUP X.X DELTA DAYS: 1  
NEXT MODE  
between  
procedures  
(1-367)  
0
0
2
ENTR EXIT  
SETUP X.X SEQ 2) DISABLED  
SETUP X.X DELTA DAYS:2  
PREV NEXT MODE  
EXIT  
<SET SET> EDIT  
EXIT  
SETUP X.X MODE: DISABLED  
SETUP X.X DELTA TIME00:00  
NEXT  
ENTR EXIT  
ENTR EXIT  
ENTR EXIT  
EXIT  
<SET SET> EDIT  
EXIT  
SETUP X.X MODE: ZERO  
Toggle keys  
to set  
delay time for  
each iteration  
of the  
sequence:  
HH:MM  
PREV NEXT  
SETUP X.X DELTA TIME: 00:00  
0
0
:3  
0
ENTR EXIT  
SETUP X.X MODE: ZERO–SPAN  
(0 – 24:00)  
PREV NEXT  
SETUP X.X DELTA TIEM:00:30  
<SET SET> EDIT  
EXIT  
SETUP X.X SEQ 2) ZERO–SPAN, 1:00:00  
PREV NEXT MODE SET  
SETUP X.X DURATION:15.0 MINUTES  
Toggle keys  
to set  
duration for  
each  
iteration of  
the  
sequence:  
Set in  
<SET SET> EDIT  
EXIT  
ENTR EXIT  
EXIT  
Default  
value is  
ON  
SETUP X.X TIMER ENABLE: ON  
SET> EDIT  
EXIT  
SETUP X.X DURATION 15.0MINUTES  
3
0
.0  
SETUP X.X STARTING DATE: 01–JAN–02  
Decimal  
minutes  
from  
<SET SET> EDIT  
EXIT  
0.1 – 60.0  
SETUP X.X DURATION:30.0 MINUTES  
Toggle keys  
to set  
day, month &  
year:  
<SET SET> EDIT  
SETUP X.X STARTING DATE: 01–JAN–02  
0
4
SEP  
0
3
ENTR EXIT  
Format :  
DD-MON-YY  
SETUP X.X  
CALIBRATE: OFF  
SETUP X.X STARTING DATE: 04–SEP–03  
<SET SET> EDIT  
EXIT  
ENTR EXIT  
EXIT  
<SET SET> EDIT  
EXIT  
Toggle key  
between  
Off and  
ON  
SETUP X.X  
ON  
CALIBRATE: OFF  
SETUP X.X STARTING DATE: 04–SEP–03  
<SET SET> EDIT  
EXIT  
SETUP X.X  
CALIBRATE: ON  
Toggle keys to  
set time:  
SETUP X.X STARTING TIME:00:00  
<SET SET> EDIT  
Format : HH:MM  
<SET SET> EDIT  
EXIT  
This is a 24 hr  
clock .  
PM hours are  
13 – 24.  
SETUP X.X SEQ 2) ZERO–SPAN, 2:00:30  
EXIT returns  
to the SETUP  
Menu  
SETUP X.X STARTING TIME:00:00  
Example  
2:15 PM = 14:15  
PREV NEXT MODE SET  
EXIT  
1
4
: 1  
5
ENTR EXIT/  
Sequence  
Delta Time  
MODE  
Delta Days  
150  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual  
Calibration Procedures  
7.6.1. AutoCal with Auto or Dual Reporting Ranges Modes Selected  
SETUP C.4  
<SET  
RANGE TO CAL: LOW  
EDIT  
EXIT  
SETUP C.4  
RANGE TO CAL: LOW  
RANGE TO CAL: HIGH  
LOW HIGH  
ENTR SETUP  
SETUP C.4  
<SET  
EDIT  
EXIT  
EXIT  
SETUP C.4 SEQ 2) ZERO–SPAN, 2:00:30  
EXIT returns to the  
PRIMARY SETUP  
Menu  
PREV NEXT MODE SET  
NOTE  
In order to automatically calibrate both the HIGH and LOW ranges, you must set up a  
separate sequence for each.  
7.7. Calibration Quality  
After completing one of the calibration procedures described above, it is important to evaluate the analyzer’s  
calibration SLOPE and OFFSET parameters. These values describe the linear response curve of the analyzer.  
The values for these terms, both individually and relative to each other, indicate the quality of the calibration. To  
perform this quality evaluation, you will need to record the values of both test functions (Section 6.2.1 or  
Appendix A-3), all of which are automatically stored in the iDAS channel CALDAT for data analysis,  
documentation and archival.  
Make sure that these parameters are within the limits listed in Table 7-3 and frequently compare them to those  
values on the Final Test and Checkout Sheet that came attached to your manual, which should not be  
significantly different. If they are, refer to the troubleshooting Chapter 11.  
Table 7-3: Calibration Data Quality Evaluation  
FUNCTION  
SLOPE  
MINIMUM VALUE  
0.700  
OPTIMUM VALUE  
1.000  
MAXIMUM VALUE  
1.300  
OFFS  
-0.500  
0.000  
0.500  
These values should not be significantly different from the values recorded on the Teledyne Instruments  
Final Test and Validation Data sheet that was shipped with your instrument. If they are, refer to the  
troubleshooting Chapter 11.  
05232 Rev B3  
151  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Calibration Procedures  
Model 360E Instruction Manual  
The default iDAS configuration records all calibration values in channel CALDAT as well as all calibration check  
(zero and span) values in its internal memory. Up to 200 data points are stored for up 4 years of data (on  
weekly calibration checks) and a lifetime history of monthly calibrations. Review these data to see if the zero  
and span responses change over time. These channels also store the STABIL value (standard deviation of CO  
concentration) to evaluate if the analyzer response has properly leveled off during the calibration procedure.  
Finally, the CALDAT channel also stores the converter efficiency for review and documentation.  
If your instrument has an O2 sensor option installed that should be calibrated as well.  
User Notes  
152  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual  
EPA Protocol Calibration  
8. EPA PROTOCOL CALIBRATION  
At the writing of this manual there is no EPA requirements for the monitoring of CO2 or published calibration  
protocols.  
User Notes  
05232 Rev B3  
153  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual  
MAINTENANCE SCHEDULE & PROCEDURES  
9. MAINTENANCE SCHEDULE & PROCEDURES  
Predictive diagnostic functions, including data acquisition records, failure warnings and test functions built into the  
analyzer, allow the user to determine when repairs are necessary without performing painstaking preventative  
maintenance procedures. There are, however, a minimal number of simple procedures that when performed  
regularly will ensure that the analyzer continues to operate accurately and reliably over its lifetime. Repairs and  
troubleshooting are covered in Chapter 11 of this manual.  
9.1. Maintenance Schedule  
Table 9-1 shows a typical maintenance schedule for the analyzer. Please note that in certain environments (i.e.  
dusty, very high ambient pollutant levels) some maintenance procedures may need to be performed more often  
than shown.  
NOTE  
A Span and Zero Calibration Check (see CAL CHECK REQ’D Column of Table 9-1) must be  
performed following certain of the maintenance procedure listed below.  
See Sections 7.3, 7.5 and 7.6 for instructions on performing checks.  
CAUTION  
Risk of electrical shock. Disconnect power before performing any of the following  
operations that require entry into the interior of the analyzer.  
NOTE  
The operations outlined in this chapter are to be performed by qualified maintenance  
personnel only.  
05232 Rev B3  
155  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
MAINTENANCE SCHEDULE & PROCEDURES  
Model 360E Instruction Manual  
9.2. Predicting Failures Using the Test Functions  
The Test Functions can be used to predict failures by looking at how their values change over time. Initially it  
may be useful to compare the state of these Test Functions to the values recorded on the printed record of the  
final calibration performed on your instrument at the factory, p/n 04307. Table 9-3 can be used as a basis for  
taking action as these values change with time. The internal data acquisition system (iDAS) is a convenient way  
to record and track these changes. Use APICOM to download and review this data from a remote location.  
Table 9-3: Predictive uses for Test Functions  
FUNCTION  
STABILITY  
CONDITION  
BEHAVIOR  
INTERPRETATION  
Pneumatic Leaks – instrument & sample system  
Detector deteriorating  
Zero Cal  
Increasing  
Source Aging  
CO2 MEAS  
Detector deteriorating  
Zero Cal  
Zero Cal  
Decreasing  
Increasing  
Optics getting dirty or contaminated  
Source Aging  
Detector deteriorating  
Contaminated zero gas (H2O)  
Source Aging  
Detector deteriorating  
GFC Wheel Leaking  
Decreasing  
Increasing  
Pneumatic Leaks  
MR RATIO  
Contaminated zero gas (CO2)  
Source Aging  
Pneumatic Leaks – instrument & sample system  
Calibration system deteriorating  
Source Aging  
Span Cal  
Sample  
GFC Wheel Leaking  
Decreasing  
Calibration system deteriorating  
Pneumatic Leak between sample inlet and Sample Cell  
Change in sampling manifold  
Dirty particulate filter  
Increasing > 1”  
PRES  
Pneumatic obstruction between sample inlet and Sample  
Cell  
Decreasing > 1”  
Obstruction in sampling manifold  
Mechanical Connection between IR-Detector and Sample  
Cell deteriorating  
Any, but with  
Bench Temp at  
48°C  
PHT DRIVE  
Increasing  
IR-Photodetector deteriorating  
See MR Ratio - Zero Cal Decreasing above  
See MR Ratio - Zero Cal Increasing above  
See MR Ratio - Span Cal Decreasing above  
See MR Ratio – Span Cal Increasing above  
Increasing  
Decreasing  
Increasing  
Decreasing  
OFFSET  
SLOPE  
Zero Cal  
Span Cal  
158  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Model 360E Instruction Manual  
MAINTENANCE SCHEDULE & PROCEDURES  
9.3. Maintenance Procedures  
The following procedures are to be performed periodically as part of the standard maintenance of the Model  
360E.  
9.3.1. Replacing the Sample Particulate Filter  
The particulate filter should be inspected often for signs of plugging or contamination. We recommend that when  
you change the filter, handle it and the wetted surfaces of the filter housing as little as possible. Do not touch  
any part of the housing, filter element, PTFE retaining ring, glass cover and the o-ring.  
To change the filter:  
1. Turn OFF the analyzer to prevent drawing debris into the instrument.  
2. Open the M360E’s hinged front panel and unscrew the knurled retaining ring on the filter assembly.  
Figure 9-1: Sample Particulate Filter Assembly  
3. Carefully remove the retaining ring, PTFE o-ring, glass filter cover and filter element.  
4. Replace the filter, being careful that the element is fully seated and centered in the bottom of the holder.  
5. Re-install the PTFE o-ring with the notches up, install the glass cover, then screw on the retaining ring  
and hand tighten. Inspect the seal between the edge of filter and the o-ring to assure a proper seal.  
6. Re-start the Analyzer.  
9.3.2. Rebuilding the Sample Pump  
The diaphragm in the sample pump periodically wears out and must be replaced. A sample rebuild kit is  
available – see Appendix B of this manual for the part number of the pump rebuild kit. Instructions and diagrams  
are included with the kit.  
Always perform a Flow and Leak Check after rebuilding the Sample Pump.  
05232 Rev B3  
159  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
MAINTENANCE SCHEDULE & PROCEDURES  
Model 360E Instruction Manual  
9.3.3. Performing Leak Checks  
Leaks are the most common cause of analyzer malfunction; Section 9.3.3.1 presents a simple leak check  
procedure. Section 9.3.3.2 details a more thorough procedure.  
9.3.3.1. Vacuum Leak Check and Pump Check  
This method is easy and fast. It detects, but does not locate most leaks; it also verifies that the sample pump is  
in good condition.  
1. Turn the analyzer ON, and allow enough time for flows to stabilize.  
2. Cap the sample inlet port.  
3. After several minutes, when the pressures have stabilized, note the following. In the TEST menu, note  
the SAMPLE PRESSURE reading.  
4. If the reading is < 10 in-Hg, the pump is in good condition and there are no large leaks.  
5. Check the sample gas flow. If the flow is <10 cm3/min and stable, there are no large leaks in the  
instrument’s pneumatics.  
9.3.3.2. Pressure Leak Check  
If you can’t locate the leak by the above procedure, use the following procedure. Obtain a leak checker similar  
to the Teledyne Instruments part number 01960, which contains a small pump, shut-off valve, and pressure  
gauge. Alternatively, a convenient source of low-pressure gas is a tank of span gas, with the two-stage regulator  
adjusted to less than 15 psi with a shutoff valve and pressure gauge.  
CAUTION  
Do not use bubble solution with vacuum applied to the analyzer. The solution may  
contaminate the instrument. Do not exceed 15 PSIG pressure.  
1. Turn OFF power to the instrument.  
2. Install a leak checker or tank of gas as described above on the sample inlet at the rear panel.  
3. Remove the instrument cover and locate the inlet side of the sample pump. Remove the flow assembly  
from the pump and plug it with the appropriate gas-tight fitting.  
4. Pressurize the instrument with the leak checker, allowing enough time to fully pressurize the instrument  
through the critical flow orifice. Check each fitting with soap bubble solution, looking for bubbles. Once  
the fittings have been wetted with soap solution, do not re-apply vacuum, as it will suck soap solution  
into the instrument and contaminate it. Do not exceed 15 psi pressure.  
5. If the instrument has one of the zero and span valve options, the normally closed ports on each valve  
should also be separately checked. Connect the leak checker to the normally closed ports and check  
with soap bubble solution.  
6. Once the leak has been located and repaired, the leak-down rate should be < 1 in-Hg (0.4 psi) in 5  
minutes after the pressure is shut off.  
160  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Model 360E Instruction Manual  
MAINTENANCE SCHEDULE & PROCEDURES  
9.3.4. Performing a Sample Flow Check  
CAUTION  
Always use a separate calibrated flow meter capable of measuring flows in the 0 – 1000  
cc/min range to measure the gas flow rate though the analyzer.  
DO NOT use the built in flow measurement viewable from the Front Panel of the  
instrument. This measurement is only for detecting major flow interruptions such as  
clogged or plugged gas lines.  
See Figure 3-2 for sample port location.  
1. Attach the Flow Meter to the sample inlet port on the rear panel. Ensure that the inlet to the Flow Meter  
is at atmospheric f.  
2. Sample flow should be 800 cc/min 10%.  
3. Once an accurate measurement has been recorded by the method described above, adjust the  
analyzer’s internal flow sensors (See Section 6.13.8)  
Low flows indicate blockage somewhere in the pneumatic pathway, typically a plugged sintered filter or critical  
flow orifice in one of the analyzer’s flow control assemblies. High flows indicate leaks downstream of the Flow  
Control Assembly.  
9.3.5. Cleaning the Optical Bench  
The M360E sensor assembly and optical bench is complex and delicate. Disassembly and cleaning is not  
recommended. Please check with the factory before disassembling the optical bench.  
9.3.6. Cleaning Exterior Surfaces of the M360E  
If necessary, the exterior surfaces of the M360E can be cleaned with a clean damp cloth. Do not submerge any  
part of the instrument in water or cleaning solution.  
User Notes  
05232 Rev B3  
161  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual  
THEORY OF OPERATION  
10. THEORY OF OPERATION  
The Model 360E Gas Filter Correlation Carbon Dioxide Analyzer is a microprocessor-controlled analyzer that  
determines the concentration of carbon dioxide (CO2) in a sample gas drawn through the instrument. It requires  
that sample and calibration gasses be supplied at ambient atmospheric pressure in order to establish a stable  
gas flow through the sample chamber where the gases ability to absorb infrared radiation is measured.  
Calibration of the instrument is performed in software and does not require physical adjustments to the  
instrument. During calibration the microprocessor measures the current state of the IR Sensor output and  
various other physical parameters of the instrument and stores them in memory.  
The microprocessor uses these calibration values, the IR absorption measurements made on the sample gas  
along with data regarding the current temperature and pressure of the gas to calculate a final co2 concentration.  
This concentration value and the original information from which it was calculated are stored in one of the unit’s  
internal data acquisition system (iDAS - See Sections 6.7) as well as reported to the user via a vacuum  
florescent display or a variety of digital and analog signal outputs.  
10.1. Measurement Method  
10.1.1. Beer’s Law  
The basic principle by which the analyzer works is called Beer’s Law. It defines the how light of a specific  
wavelength is absorbed by a particular gas molecule over a certain distance. The mathematical relationship  
between these three parameters is:  
I = Io e-αLc  
Where:  
Io  
is the intensity of the light if there was no absorption.  
is the intensity with absorption.  
is the absorption path, or the distance the light travels as it is being absorbed.  
I
L
C
is the concentration of the absorbing gas. In the case of the Model 360E, carbon dioxide (CO2).  
α
is the absorption coefficient that tells how well CO2 absorbs light at the specific wavelength of  
interest.  
10.1.2. Measurement Fundamentals  
In the most basic terms, the Model 360E uses a high energy heated element to generate a beam of broad-band  
IR light with a known intensity (measured during Instrument calibration. This beam is directed through  
multi-pass cell filled with sample gas. The sample cell uses mirrors at each end to reflect the IR beam back and  
forth through the sample gas to generate a 2.5 meter absorption path (see Figure 10–1). This length was  
chosen to give the analyzer maximum sensitivity to fluctuations in CO2 density.  
05232 Rev B3  
163  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
THEORY OF OPERATION  
Model 360E Instruction Manual  
Band-Pass Filter  
Sample Chamber  
IR  
Source  
Photo-Detector  
IR Beam  
Figure 10-1: Measurement Fundamentals  
Upon exiting the sample cell, the beam shines through a band-pass filter that allows only light at a wavelength of  
4.3 µm to pass. Finally, the beam strikes a solid-state photo-detector that converts the light signal into a  
modulated voltage signal representing the attenuated intensity of the beam.  
10.1.3. Gas Filter Correlation  
Unfortunately, water vapor absorbs light at 4.3 µm too. To overcome the interfering effects of water vapor the  
Model 360E adds another component to the IR light path called a gas filter correlation (GFC) wheel (see Figure  
10-2).  
Measurement Cell  
(Pure N2)  
Reference Cell  
(N2 with CO2)  
Figure 10-2: GFC Wheel  
10.1.3.1. The GFC Wheel  
A GFC wheel is a metallic wheel into which two chambers are carved. The chambers are sealed on both sides  
with material transparent to 4.3 µm IR radiation creating two airtight cavities. Each cavity is filled with specially  
composed gases. One cell is filled with pure N2 (the measure cell). The other is filled with a combination of N2  
and a high concentration of CO2 (the reference cell).  
164  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Model 360E Instruction Manual  
THEORY OF OPERATION  
IR unaffected by N2 in Measurement Cell  
Δ H  
IR IS affected by CO in Reference Cell  
M
IR  
Source  
Photo-Detector  
R
GFC Wheel  
Figure 10-3: Measurement Fundamentals with GFC Wheel  
As the GFC wheel spins, the IR light alternately passes through the two cavities. When the beam is exposed to  
the reference cell, the CO2 in the gas filter wheel strips the beam of most of the IR at 4.3μm. When the light  
beam is exposed to the measurement cell, the N2 in the filter wheel does not absorb IR light. This results in a  
fluctuation in the intensity of the IR light striking the photo-detector (See Figure 10-3) that results in the output of  
the detector resembling a square wave.  
10.1.3.2. The Measure Reference Ratio  
The Model 360E determines the amount of CO2 in the sample chamber by computing the ratio between the peak  
of the measurement pulse (CO2 MEAS) and the peak of the reference pulse (CO2 REF).  
If no gases exist in the sample chamber that absorb light at 4.3μm, the high concentration of CO2 in the gas  
mixture of the reference cell will attenuate the intensity of the IR beam by 60% giving a M/R ratio of  
approximately 2.4:1.  
Adding CO2 to the sample chamber causes the peaks corresponding to both cells to be attenuated by a further  
percentage. Since the intensity of the light passing through the measurement cell is greater, the effect of this  
additional attenuation is greater. This causes CO2 MEAS to be more sensitive to the presence of CO2 in the  
sample chamber than CO2 REF and the ratio between them (M/R) to move closer to 1:1 as the concentration of  
CO2 in the sample chamber increases.  
05232 Rev B3  
165  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
THEORY OF OPERATION  
Model 360E Instruction Manual  
IR unaffected by N2 in Measurement Cell of  
the GDC Wheel and no additional CO2 in  
the Sample Chamber  
CO2 MEAS  
CO2 REF  
IR affected by CO2 in Reference Cell  
with no interfering gas in the Sample  
Chamber  
IR shinning through Measurement Cell of  
the GDC Wheel is reduced by additional  
CO2 in the Sample Chamber  
M/R  
is reduced  
IR shining through Reference Cell is  
also reduced by additional CO2 in the  
Sample Chamber, but to a lesser extent  
Figure 10-4: Affect of CO2 in the Sample on CO2 MEAS & CO2 REF  
Once the Model 360E has computed this ratio, a look-up table is used, with interpolation, to linearize the  
response of the instrument. This linearized concentration value is combined with calibration SLOPE and  
OFFSET values to produce the CO2 concentration which is then normalized for changes in sample pressure.  
10.1.4. Interference and Signal to Noise Rejection  
If an interfering gas, such as H2O vapor is introduced into the sample chamber, the spectrum of the IR beam is  
changed in a way that is identical for both the reference and the measurement cells, but without changing the  
ratio between the peak heights of CO2 MEAS and CO2 REF. In effect, the difference between the peak heights  
remains the same.  
M/R  
is Shifted  
IR shining through both cells is effected  
equally by interfering gas in the Sample  
Chamber  
Figure 10-5: Effects of Interfering Gas on CO2 MEAS & CO2 REF  
166  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Model 360E Instruction Manual  
THEORY OF OPERATION  
Thus, the difference in the peak heights and the resulting M/R ratio is only due to CO2 and not to interfering  
gases. In this way, Gas filter correlation rejects the effects of interfering gases and so that the analyzer  
responds only to the presence of CO2.  
To improve the signal-to-noise performance of the IR photo-detector, the GFC wheel also incorporates an optical  
mask that chops the IR beam into alternating pulses of light and dark at six times the frequency of the  
measure/reference signal. This limits the detection bandwidth helping to reject interfering signals from outside  
this bandwidth improving the signal to noise ratio.  
The IR Signal as the Photo-Detector sees it  
after being chopped by the GFC Wheel  
Screen  
CO2 MEAS  
CO2 REF  
Figure 10-6: Chopped IR Signal  
10.1.4.1. Ambient CO2 Interference Rejection  
CO2 absorbs IR light very well. So well that even the narrow volume of ambient air between the IR source and  
the sample chamber is enough to alter the analyzer’s measured concentration of CO2. Also, ambient air, which  
averages around 350 ppm to 400 ppm, will vary significantly over the course of the day. The ambient CO2  
concentration can rise as high as 1 000 ppm during the time of the day when people are present. It can fluctuate  
300 ppm as the photosynthesis of plant life in the nearby area increases during the day and decreases at  
night.  
The basic design of the M360E rejects most of this interference at a 100:1 ratio; however this still can allow  
small fluctuations in CO2 concentration during the course of the day. To completely remove all effects of  
ambient CO2 from the analyzer’s measurement of CO2, dried air, scrubbed of all CO2 is pumped into the GFC  
wheel housing to purge all ambient CO2 (see Figure 10-7)  
05232 Rev B3  
167  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
THEORY OF OPERATION  
Model 360E Instruction Manual  
10.2. Pneumatic Operation  
CAUTION  
It is important that the sample airflow system is both leak tight and not pressurized  
over ambient pressure.  
Regular leak checks should be performed on the analyzer as described in the  
maintenance schedule, Table 9-1.  
Procedures for correctly performing leak checks can be found in Section 9.3.3.  
An internal pump evacuates the sample chamber creating a small vacuum that draws sample gas into the  
analyzer. Normally the analyzer is operated with its inlet near ambient pressure either because the sample is  
directly drawn at the inlet or a small vent is installed at the inlet. There are several advantages to this “pull  
through” configuration.  
By placing the pump down stream from the sample chamber several problems are avoided.  
First the pumping process heats and compresses the sample air complicating the measurement  
process.  
Additionally, certain physical parts of the pump itself are made of materials that might chemically react  
with the sample gas.  
Finally, in certain applications where the concentration of the target gas might be high enough to be  
hazardous, maintaining a negative gas pressure relative to ambient means that should a minor leak  
occur, no sample gas will be pumped into the atmosphere surrounding analyzer.  
10.2.1. Sample Gas Flow  
SAMPLE GAS  
INLET  
EXHAUST  
GAS OUTLET  
PUMP  
SAMPLE  
PRESSURE  
SENSOR  
Flow / Pressure  
Sensor PCA  
O3 FLOW  
SENSOR  
Sample Gas  
Flow Control  
PURGE GAS  
INLET  
Purge Gas  
Pressure Control  
Figure 10-7: Internal Pneumatic Flow – Basic Configuration  
168  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Model 360E Instruction Manual  
THEORY OF OPERATION  
10.2.2. Flow Rate Control  
To maintain a constant flow rate of the sample gas through the instrument, the Model 360E uses special flow  
control assemblies located in the purge gas line at the entrance to the GFC wheel housing and in the exhaust  
gas line just before the pump (see Figure 10-7). These assemblies consists of:  
A critical flow orifice.  
Two o-rings: Located just before and after the critical flow orifice, the o-rings seal the gap between the  
walls of assembly housing and the critical flow orifice.  
A spring: Applies mechanical force needed to form the seal between the o-rings, the critical flow orifice  
and the assembly housing.  
10.2.2.1. Critical Flow Orifice  
The most important component of this flow control assembly is the critical flow orifice.  
Critical flow orifices are a remarkably simple way to regulate stable gas flow rates. They operate without moving  
parts by taking advantage of the laws of fluid dynamics. By restricting the flow of gas though the orifice, a  
pressure differential is created. This pressure differential combined with the action of the analyzer’s pump draws  
the gas through the orifice.  
As the pressure on the downstream side of the orifice (the pump side) continues to drop, the speed that the gas  
flows though the orifice continues to rise. Once the ratio of upstream pressure to downstream pressure is  
greater than 2:1, the velocity of the gas through the orifice reaches the speed of sound. As long as that ratio  
stays at least 2:1 the gas flow rate is unaffected by any fluctuations, surges, or changes in downstream pressure  
because such variations only travel at the speed of sound themselves and are therefore cancelled out by the  
sonic shockwave at the downstream exit of the critical flow orifice.  
CRITICAL  
FLOW  
ORIFICE  
AREA OF  
LOW  
AREA OF  
HIGH  
PRESSURE  
PRESSURE  
Sonic  
Shockwave  
O-RINGS  
SPRING  
FILTER  
Figure 10-8: Flow Control Assembly & Critical Flow Orifice  
The actual flow rate of gas through the orifice (volume of gas per unit of time), depends on the size and shape of  
the aperture in the orifice. The larger the hole, the more gas molecules, moving at the speed of sound, pass  
through the orifice. Because the flow rate of gas through the orifice is only related to the minimum 2:1 pressure  
differential and not absolute pressure the flow rate of the gas is also unaffected by degradations in pump  
efficiency due to age.  
The critical flow orifice used in the Model 360E is designed to provide a flow rate of 800 cm3/min.  
05232 Rev B3  
169  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
THEORY OF OPERATION  
Model 360E Instruction Manual  
10.2.3. Purge Gas Pressure Control  
In order to ensure that all of the ambient CO2 is purged from the GFC Wheel housing an adequate supply of  
dried air, scrubbed of CO2 must be supplied to the PURGE AIR inlet at the back of the instrument.  
The minimum gas pressure of the source of purge air should be 7.5 psig.  
If the source of the purge air is shared by a Teledyne Instruments M700 (as shown in figure 3-7) the  
minimum gas pressure should be 25 psig and should not exceed 35 psig.  
In order to maintain the proper pressure differential between the inside of the GFC wheel housing and ambient  
air, the M360 design includes a manually settable pressure regulator that maintains the pressure of the purge air  
feed at 7.5 psig.  
10.2.4. Particulate Filter  
The Model 360E Analyzer comes equipped with a 47 mm diameter, Teflon, particulate filter with a 5 micron pore  
size. The filter is accessible through the front panel, which folds down to allow access, and should be changed  
according to the suggested maintenance schedule described in Table 9-1.  
10.2.5. Pneumatic Sensors  
10.2.5.1. Sample Pressure Sensor  
An absolute value pressure transducer plumbed to the outlet of the sample chamber is used to measure sample  
pressure. The output of the sensor is used to compensate the concentration measurement for changes in air  
pressure. This sensor is mounted to a printed circuit board with the sample flow sensor on the sample chamber;  
see the following section and Figure 3-3.  
10.2.5.2. Sample Flow Sensor  
A thermal-mass flow sensor is used to measure the sample flow through the analyzer. The sensor is calibrated  
at the factory with ambient air or N2, but can be calibrated to operate with samples consisting of other gases  
such as CO2, See Section 9.3.4. This sensor is mounted to a printed circuit board with the Sample Pressure  
sensor on the sample chamber; see the previous section and Figure 3-3.  
170  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Model 360E Instruction Manual  
10.3. Electronic Operation  
10.3.1. Overview  
THEORY OF OPERATION  
Figure 10-9 shows a block diagram of the major electronic components of the Model 360E.  
At its heart the analyzer is a microcomputer (CPU) that controls various internal processes, interprets data,  
makes calculations, and reports results using specialized firmware developed by Teledyne Instruments. It  
communicates with the user as well as receives data from and issues commands to a variety of peripheral  
devices via a separate printed circuit assembly called the Mother Board.  
The mother board collects data, performs signal conditioning duties and routs incoming and outgoing signals  
between the CPU and the analyzer’s other major components.  
Data is generated by a gas-filter-correlation optical bench which outputs an analog signal corresponding to the  
concentration of CO2 in the sample gas. This analog signal is transformed into two, pre-amplified, DC voltages  
(CO2 MEAS and CO2 REF) by a synchronous demodulator printed circuit assembly. CO2 MEAS and CO2 REF  
are converted into digital data by a unipolar, analog-to-digital converter, located on the mother board.  
A variety of sensors report the physical and operational status of the analyzer’s major components, again  
through the signal processing capabilities of the mother board. These status reports are used as data for the  
CO2 concentration calculation and as trigger events for certain control commands issued by the CPU. They are  
stored in memory by the CPU and in most cases can be viewed but the user via the front panel display.  
The CPU communicates with the user and the outside world in a variety of manners:  
Through the analyzer’s keyboard and vacuum florescent display over a clocked, digital, serial I/O bus  
(using a protocol called I2C);  
RS 232 & RS485 Serial I/O channels;  
Via an optional Ethernet communications card:  
Various DCV and DCA analog outputs, and  
Several sets of Digital I/O channels.  
Finally, the CPU issues commands via a series of relays and switches (also over the I2C bus) located on a  
separate printed circuit assembly to control the function of key electromechanical devices such as heaters,  
motors and valves.  
05232 Rev B3  
171  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
THEORY OF OPERATION  
Model 360E Instruction Manual  
Analog Outputs  
A1  
A2  
Back Panel  
Connectors  
Optional  
4-20 mA  
Control Inputs:  
1 – 8  
COM1 COM2  
A3  
A4  
Status Outputs:  
1 – 6  
Optional  
Ethernet  
Interface  
Analog  
Outputs  
(D/A)  
External  
Digital I/O)  
PC 104  
CPU Card  
A/D  
Converter  
(V/F)  
RS–232  
or RS-485  
Power-Up  
Circuit  
Disk On  
Chip  
RS – 232  
MOTHER  
BOARD  
Flash Chip  
Box  
Temp  
PC 104 Bus  
Zero/Span  
Valve  
Options  
PUMP  
Thermistor  
Interface  
Internal  
Digital I/O  
I2C  
Bus  
Sensor Inputs  
SAMPLE  
TEMP  
C
O
C
O
RELAY  
BOARD  
Sample Flow  
& Pressure  
Sensors  
Optional  
O2 Sensor  
Keyboard &  
Display  
CPU Status  
M
E
A
S
R
E
F
LED  
BENCH  
TEMP  
TEC Control  
IR  
Source  
PHT  
WHEEL  
TEMP  
Photo-  
detector  
SYNC  
DEMOD  
Drive  
Detector  
Output  
GFC  
Motor  
GFC  
Wheel  
O2 SENSOR  
TEMP  
(optional)  
Optical  
Bench  
Schmidt  
Trigger  
Wheel  
Heater  
Segment Sensor  
Bench Heater  
M / R Sensor  
Figure 10-9: 360E Electronic Block Diagram  
172  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 360E Instruction Manual  
THEORY OF OPERATION  
10.3.2. CPU  
The Model 360E’s CPU is a, low power (5 VDC, 0.8A max), high performance, 386-based microcomputer  
running MS-DOS. Its operation and assembly conform to the PC/104 Specification version 2.3 for embedded  
PC and PC/AT applications. It has 2 MB of DRAM on board and operates at 40MHz over an internal 32-bit data  
and address bus. Chip to chip data handling is performed by two 4-channel DMA devices over data busses of  
either 8-bit or 16-bit configuration. The CPU supports both RS-232 and RS-485 serial I/O.  
The CPU includes two types of non-volatile data storage.  
Disk-On-Chip  
While technically an EEPROM, the Disk –on-Chip (DOC), this device appears to the CPU as, behaves as, and  
performs the same function in the system as an 8MB disk drive. It is used to store the operating system for the  
computer, the Teledyne Instruments Firmware, and most of the operational data generated by the analyzer’s  
internal data acquisition system (iDAS - See Section 6.7).  
Flash Chip  
Another, smaller EEPROM used to store critical calibration and configuration data. Segregating this data on a  
separate, less heavily accessed chip significantly decreases the chance of this key data being corrupted.  
10.3.3. Optical Bench & GFC Wheel  
Electronically, the Model 360E’s optical bench, GFC wheel and associated components do more than simply  
measure the amount of CO2 present in the sample chamber. A variety of other critical functions are performed  
here as well.  
10.3.3.1. Temperature Control  
Because the temperature of a gas affects its density and therefore the amount of light absorbed by that gas it is  
important to reduce the effect of fluctuations in ambient temperature on the Model 360E’s measurement of CO2.  
To accomplish this both the temperature of the sample chamber and the GFC Wheel are maintained at constant  
temperatures above their normal operating ranges.  
Bench Temperature: To minimize the effects of ambient temperature variations on the sample measurement, the  
sample chamber is heated to 48C (8 degrees above the maximum suggested ambient operating temperature  
for the analyzer). A strip heater attached to the underside of the chamber housing is the heat source. The  
temperature of the sample chamber is sensed by a thermistor, also attached to the sample chamber housing.  
Wheel Temperature: To minimize the effects of temperature variations caused by the near proximity of the IR  
Source to the GFC wheel on the gases contained in the wheel, it is also raised to a high temperature level.  
Because the IR Source itself is very hot, the set point for this heat circuit is 68C. A cartridge heater implanted  
into the heat sync on the motor is the heat source. The temperature of the wheel/motor assembly is sensed by a  
thermistor also inserted into the heat sync.  
Both heaters operate off of the AC line voltage supplied to the instrument.  
10.3.3.2. IR Source  
The light used to detect CO2 in the sample chamber is generated by an element heated to approximately 1100oC  
producing infrared radiation across a broad band. This radiation is optically filtered after it has passed through  
the GFC Wheel and the sample chamber and just before it reaches the photo-detector to eliminate all black body  
radiation and other extraneous IR emitted by the various components of those components.  
05232 Rev B3  
173  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
THEORY OF OPERATION  
Model 360E Instruction Manual  
10.3.3.3. GFC Wheel  
A synchronous AC motor turns the GFC wheel motor. For analyzers operating on 60Hz line power this motor  
turns at 1800 rpm. For those operating on 50Hz line power the spin rate is 1500 rpm. The actual spin rate is  
unimportant within a large rate since a phase lock loop circuit is used to generate timing pulses for signal  
processing.  
In order to accurately interpret the fluctuations of the IR beam after it has passed through the sample gas, the  
GFC wheel several other timing signals are produced by other photo emitters/detectors. These devices consist  
of a combination LED and detector mounted so that the light emitted by the LED shines through the same mask  
on the GFC wheel that chops the IR beam.  
KEY:  
Detection Beam shining  
through MEASUREMENT  
side of GFC Wheel  
Detection Beam shining  
through REFERENCE side  
of GFC Wheel  
IR Detection Ring  
Segment Sensor Ring  
M/R Sensor Ring  
Figure 10-10: GFC Light Mask  
M/R Sensor  
This emitter/detector assembly that produces a signal that shines through a portion of the mask that allows light  
to pass for half of a full revolution of the wheel. The resulting light signal tells the analyzer whether the IR beam  
is shining through the measurement or the reference side of the GFC wheel.  
Segment Sensor  
Light from this emitter/detector pair shines through a portion of the mask that is divided into the same number of  
segments as the IR detector ring. It is used by the synchronous / demodulation circuitry of the analyzer to latch  
onto the most stable part of each measurement and reference IR pulse.  
174  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Model 360E Instruction Manual  
THEORY OF OPERATION  
Measurement  
Pulses  
Reference  
Pulses  
IR Beam  
Pulses  
Segment Sensor  
Pulses  
MR Sensor  
Pulses  
Figure 10-11: Segment Sensor and M/R Sensor Output  
Schmidt Triggers  
To ensure that the waveforms produced by the Segment Sensor and the M/R Sensor are properly shaped and  
clean, these signals are passed through a set of Schmidt Triggers circuits.  
10.3.3.4. IR Photo-Detector  
The IR beam is converted into an electrical signal by a cooled solid-state photo-conductive detector. The  
detector is composed of a narrow-band optical filter, a piece of lead-salt crystal whose electrical resistance  
changes with temperature, and a two-stage thermo-electric cooler.  
When the analyzer is on, a constant electrical current is directed through the detector, The IR beam is focused  
onto the detector surface, raising its temperature and lowering its electrical resistance that results in a change in  
the voltage drop across the detector.  
During those times that the IR beam is bright, the temperature of the detector is high; the resistance of the  
detector is correspondingly low and the its output voltage output is low. During those times when the IR beam  
intensity is low or completely blocked by the GFC Wheel mask, the temperature of the detector is lowered by the  
two-stage thermo-electric cooler, increasing the detectors resistance and raising the output voltage.  
10.3.4. Synchronous Demodulator (Sync/Demod) Assembly  
10.3.4.1. Overview  
While the photo-detector converts fluctuations of the IR beam into electronic signals, the Sync / Demod Board  
amplifies these signals and converts them into usable information. Initially the output by the photo-detector is a  
complex and continuously changing waveform made up of Measure and Reference pulses. The sync/demod  
board demodulates this waveform and outputs two analog DC voltage signals, corresponding to the peak values  
of these pulses. CO2 MEAS and CO2 REF are converted into digital signals by circuitry on the motherboard  
then used by the CPU to calculate the CO2 concentration of the sample gas.  
Additionally the synch/demod board contains circuitry that controls the photo-detector’s thermoelectric cooler as  
well as circuitry for performing certain diagnostic tests on the analyzer.  
05232 Rev B3  
175  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
THEORY OF OPERATION  
Model 360E Instruction Manual  
56V  
Bias  
CO2 MEAS  
Variable  
Gain Amp  
Sample &  
Dark  
Switch  
Pre Amp  
Photo-  
detector  
Signal  
Conditioner  
Hold  
Circuits  
TEC Control  
PHT DRIVE  
E-Test  
Generator  
CO2 Reference  
Signal  
(x4)  
Conditioner  
Thermo-Electric  
Cooler  
Control Circuit  
E Test A Gate  
E Test B Gate  
Dark Test Gate  
Measure Gate  
Compact  
Programmable  
Logic Device  
Measure Dark Gate  
Reference Gate  
Reference Dark Gate  
Phase Lock Warning  
M/R Sensor  
From GFC  
Wheel  
Segment  
Sensor  
Segment Clock  
X1 Reference  
E Test Control  
Phase  
Lock  
Loop  
x10  
From CPU  
via Mother  
Board  
10  
Dark Switch  
Control  
X10 Clock  
M/R  
Segment  
Status LED  
Status LED  
Phase Lock  
Figure 10-12: 360E Sync / Demod Block Diagram  
10.3.4.2. Signal Synchronization and Demodulation  
The signal emitted by the IR photo-detector goes through several stages of amplification before it can be  
accurately demodulated. The first is a pre-amplification stage that raises the signal to levels readable by the rest  
of the synch/demod board circuitry. The second is a variable amplification stage that is adjusted at the factory to  
compensate for performance variations of mirrors, detectors, and other components of the optical bench from  
instrument to instrument.  
The workhorses of the sync/demod board are the four sample-and-hold circuits that capture various voltage  
levels found in the amplified detector signal needed to determine the value of CO2 MEAS and CO2 REF. They  
are activated by logic signals under the control of a compact programmable logic device (PLD), which in turn  
responds to the output of the Segment Sensor and M/R Sensor described in Figure 10–11.  
176  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Model 360E Instruction Manual  
THEORY OF OPERATION  
The four sample and hold circuits are:  
DESIGNATION  
ACTIVE WHEN:  
IR BEAM PASSING THROUGH  
MEASUREMENT cell of GFC Wheel  
MEASUREMENT Cell of GFC Wheel  
REFERENCE cell of GFC Wheel  
REFERENCE cell of GFC Wheel  
SEGMENT SENSOR PULSE IS:  
Measure Gate  
Measure Dark Gate  
Reference Gate  
HIGH  
LOW  
HIGH  
LOW  
Reference Dark Gate  
Timing for activating the Sample and Hold circuits is provided by a phase lock loop circuit (PLL). Using the  
segment sensor output as a reference signal the PLL generates clock signal at ten times that frequency. This  
faster clock signal is used by the PLD to make the sample and hold circuits capture the signal during the center  
portions of the detected waveform, ignore the rising and falling edges of the detector signal.  
Sample & Hold  
Active  
Detector  
Output  
Sample & Hold  
Inactive  
Figure 10-13: Sample & Hold Timing  
10.3.4.3. Sync/Demod Status LED’s  
The following two status LED’s located on the synch/demod board provide additional diagnostic tools for  
checking the GFC wheel rotation.  
Table 10-1: Sync/Demod Status LED Activity  
LED  
D1  
FUNCTION  
STATUS OK  
FAULT STATUS  
M/R Sensor Status  
LED flashes approximately  
2/second  
LED is stuck  
ON or OFF  
D2  
Segment Sensor  
Status  
LED flashes approximately  
6/second  
LED is stuck  
ON or OFF  
See Section 11.1.4 for more information.  
05232 Rev B3  
177  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
THEORY OF OPERATION  
Model 360E Instruction Manual  
10.3.4.4. Photo-Detector Temperature Control  
The synch/demod board also contains circuitry that controls the IR photo-detector’s thermoelectric coolers. A  
drive voltage, PHT DRIVE, is supplied to the coolers by the synch/demod board which is adjusted by the  
synch/demod board based on a return signal called TEC control which alerts informs the synch/demod board of  
the detector’s temperature. The warmer the detector, the harder the coolers are driven.  
PHT DRIVE is one of the Test Functions viewable by the user via the front panel. Press <TST or TST> until it  
appears on the display.  
10.3.4.5. Dark Calibration Switch  
This switch initiates the Dark Calibration procedure. When initiated by the user (See Section 6.13.6 for more  
details), the dark calibration process opens this switch, interrupting the signal from the IR photo-detector. This  
allows the analyzer to measure any offset caused by the synch/demod board circuitry.  
10.3.4.6. Electric Test Switch  
When active this circuit generates a specific waveform intended to simulate the function of the IR photo-detector  
but with a known set of value which is substituted for the detector’s actual signal via the dark switch. It may also  
be initiated by the user (See Section 6.13.5 for more details).  
10.3.5. Relay Board  
By actuating various switches and relays located on this board, the CPU controls the status of other key  
components. The relay board receives instructions in the form of digital signals over the I2C bus, interprets  
these digital instructions and activates its various switches and relays appropriately.  
Heater Control  
The two heaters attached to the sample chamber housing and the GFC wheel motor are controlled by solid state  
relays located on the relay board.  
The GFC wheel heater is simply turned on or off, however control of the bench heater also includes circuitry that  
selects which one of its two separate heating elements is activated depending on whether the instrument is  
running on 100 VAC, 115 VAC or 230 VAC line power.  
GFC Wheel Motor Control:  
The GFC wheel operates from an AC voltage supplied by a multi-input transformer located on the relay board.  
The step-down ratio of this transformer is controlled by factory-installed jumpers to adjust for 100 VAC, 115 VAC  
or 230 VAC line power. Other circuitry slightly alters the phase of the AC power supplied to the motor during  
start up based on whether line power is 50Hz or 60 Hz.  
Normally, the GFC Wheel Motor is always turning while the analyzer is on. A physical switch located on the  
relay board can be used to turn the motor off for certain diagnostic procedures.  
178  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Model 360E Instruction Manual  
THEORY OF OPERATION  
Zero/Span Valve Options  
Any zero/span/shutoff valve options installed in the analyzer are controlled by a set of electronic switches  
located on the relay board. These switches, under CPU control, supply the +12VDC needed to activate each  
valve’s solenoid.  
IR Source  
The Relay board supplies a constant 11.5VDC to the IR Source. Under normal operation the IR source is  
always on.  
10.3.5.1. Status LED’s  
Eight LED’s are located on the analyzer’s relay board to show the current status on the various control functions  
performed by the relay board (see Figure 10-14). They are:  
Table 10-2: Relay Board Status LED’s  
LED  
COLOR  
FUNCTION  
STATUS WHEN LIT  
STATUS WHEN UNLIT  
D1  
RED  
Watchdog Circuit  
Cycles On/Off Every 3 Seconds under direct control of the  
analyzer’s CPU.  
D2  
D3  
D4  
D5  
YELLOW  
YELLOW  
YELLOW  
GREEN  
Wheel Heater  
Bench Heater  
Spare  
HEATING  
HEATING  
N/A  
NOT HEATING  
NOT HEATING  
N/A  
Sample/Cal Gas  
Valve Option  
Valve Open to CAL GAS  
FLOW  
Valve Open to SAMPLE GAS  
FLOW  
D6  
D7  
D8  
GREEN  
GREEN  
GREEN  
Zero/Span Gas  
Valve Option  
Valve Open to SPAN GAS  
FLOW  
Valve Open to ZERO GAS FLOW  
Shutoff Valve  
Option  
Valve Open to CAL GAS  
FLOW  
Valve CLOSED to CAL GAS  
FLOW  
IR SOURCE  
Source ON  
Source OFF  
DC VOLTAGE TEST  
POINTS  
STATUS LED’s  
RELAY PCA  
PN 04135  
05232 Rev B3  
179  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
THEORY OF OPERATION  
Model 360E Instruction Manual  
Figure 10-14: Location of relay board Status LED’s  
10.3.5.2. I2C Watch Dog Circuitry  
Special circuitry on the relay board monitors the activity on the I2C bus and drives LED D1. Should this LED  
ever stay ON or OFF for 30 seconds, the watchdog circuit will automatically shut of all valves as well as turn off  
the IR Source and all heaters. The GFC wheel motor will still be running as will the Sample Pump, which is not  
controlled by the relay board.  
10.3.6. Mother Board  
This printed circuit assembly provides a multitude of functions including, A/D conversion, digital input/output, PC-  
104 to I2C translation, temperature sensor signal processing and is a pass through for the RS-232 and RS-485  
signals.  
10.3.6.1. A to D Conversion  
Analog signals, such as the voltages received from the analyzer’s various sensors, are converted into digital  
signals that the CPU can understand and manipulate by the analog to digital converter (A/D). Under the control  
of the CPU, this functional block selects a particular signal input (e.g. BOX TEMP, CO2 MEAS, CO2 REF, etc.)  
and then coverts the selected voltage into a digital word.  
The A/D consists of a voltage-to-frequency (V-F) converter, a programmable logic device (PLD), three  
multiplexers, several amplifiers and some other associated devices. The V-F converter produces a frequency  
proportional to its input voltage. The PLD counts the output of the V-F during a specified time period, and sends  
the result of that count, in the form of a binary number, to the CPU.  
The A/D can be configured for several different input modes and ranges but in the M360E is used in uni-polar  
mode with a +5 V full scale. The converter includes a 1% over and under-range. This allows signals from –0.05  
V to +5.05 V to be fully converted.  
For calibration purposes, two reference voltages are supplied to the A/D converter: Reference Ground and  
+4.096 VDC. During calibration, the device measures these two voltages, outputs their digital equivalent to the  
CPU. The CPU uses these values to compute the converter’s offset and slope and uses these factors for  
subsequent conversions.  
See Section 6.13.4 for instructions on performing this calibration.  
10.3.6.2. Sensor Inputs  
The key analog sensor signals are coupled to the A/D through the master multiplexer from two connectors on  
the motherboard. 100K terminating resistors on each of the inputs prevent cross talk from appearing on the  
sensor signals.  
Co2 Measure And Reference  
These are the primary signals that are used in the computation of the CO2 concentration. They are the  
demodulated IR-sensor signals from the sync demodulator board.  
Sample Pressure And Flow  
These are analog signals from two sensors that measure the pressure and flow rate of the gas stream at the  
outlet of the sample chamber. This information is used in two ways. First, the sample pressure is used by the  
CPU to calculate CO2 Concentration. Second, the pressure and flow rate are monitored as a test function to  
assist the user in predicting and troubleshooting failures.  
180  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Model 360E Instruction Manual  
THEORY OF OPERATION  
10.3.6.3. Thermistor Interface  
This circuit provides excitation, termination and signal selection for several negative-coefficient, thermistor  
temperature sensors located inside the analyzer. They are:  
Sample Temperature Sensor  
The source of this signal is a thermistor located inside the sample chamber of the Optical Bench. It measures  
the temperature of the sample gas in the chamber. This data is used to during the calculation of the CO2  
concentration value.  
Bench Temperature Sensor  
This thermistor, attached to the sample chamber housing, reports the current temperature of the chamber  
housing to the CPU as part of the bench heater control loop.  
Wheel Temperature Sensor  
This thermistor (attached to the heat-sync on the GFC wheel motor assembly) reports the current temperature of  
the wheel/motor assembly to the CPU as part of the Wheel Heater control loop.  
Box Temperature Sensor  
A thermistor is attached to the motherboard. It measures the analyzer’s inside temperature. This information is  
stored by the CPU and can be viewed by the user for troubleshooting purposes via the front panel display (See  
Section 11.1.2).  
10.3.6.4. Analog Outputs  
The analyzer comes equipped with four analog outputs: A1, A2, A3 and A4. . In its standard configuration, the  
analyzer comes with all four of these channels set up to output a DC voltage. However, 4-20mA current loop  
drivers can be purchased for the first three of these outputs: A2, A2 & A3.  
A2 and A1 Output  
The first two, A2 and A1 are normally set up to operate in parallel so that the same data can be sent to two  
different recording devices. While the names imply that one should be used for sending data to a chart recorder  
and the other for interfacing with a data logger, either can be used for both applications.  
Both of these channels output a signal that is proportional to the CO2 concentration of the sample gas. The A1  
and A2 outputs can be slaved together or set up to operated independently. A variety of scaling factors are  
available, See Section 6.13.4 for information on setting the range type and scaling factors for these output  
channels.  
A3 Output  
Analog output channel A3 is only active when the O2 sensor option is installed in the M360E. In this case, the  
currently measured O2 concentration is output.  
Test Function Output  
The fourth analog output, labeled A4 is special. It can be set by the user (See Section 6.9.9) to carry the current  
signal level of any one of the parameters accessible through the SETUP MORE DIAG TEST CHAN  
OUTPUT submenu (See Section 6.13.9) of the unit’s software.  
Output Loop-back  
05232 Rev B3  
181  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
THEORY OF OPERATION  
Model 360E Instruction Manual  
All four analog outputs are connected back to the A/D converter through a Loop-back circuit. This permits the  
voltage outputs to be calibrated by the CPU without need for any additional tools or fixtures.  
10.3.6.5. Internal Digital I/O  
This channel is used to communicate digital status and control signals about the operation of key components of  
the Optical Bench. The CPU sends signals to the synch/demod board that initiate the ELECTRICAL TEST and  
DARK CALIBRATION procedures. Likewise, the synch/demod board uses this interface to send the SYNC  
warning signal to the CPU (See Sections 6.9.5, 6.9.6 and 11.1.1).  
10.3.6.6. External Digital I/O  
This External Digital I/O performs two functions.  
Status Outputs  
Logic-Level voltages are output through an optically isolated 8-pin connector located on the rear panel of the  
analyzer. These outputs convey good/bad and on/off information about certain analyzer conditions. They can  
be used to interface with certain types of programmable devices (See Section 6.13.1.1).  
Control Inputs  
By applying +5VDC power supplied from an external source such as a PLC or Data logger (See Section  
6.13.1.2), Zero and Span calibrations can be initiated by contact closures on the rear panel.  
10.3.7. I2C Data Bus  
An I2C data bus is used to communicate data and commands between the CPU and the keyboard/display  
interface and the relay board. I2C is a two-wire, clocked, digital serial I/O bus that is used widely in commercial  
and consumer electronic systems. A transceiver on the motherboard converts data and control signals from the  
PC-104 bus to I2C. The data is then fed to the keyboard/display interface and finally onto the relay board.  
Interface circuits on the keyboard/display interface and relay boards convert the i2c data to parallel inputs and  
outputs. An additional, interrupt line from the keyboard to the motherboard allows the CPU to recognize and  
service key presses on the keyboard.  
Power up Circuit  
This circuit monitors the +5V power supply during start-up and sets the Analog outputs, external digital I/O ports,  
and I2C circuitry to specific values until the CPU boots and the instrument software can establish control.  
182  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Model 360E Instruction Manual  
THEORY OF OPERATION  
10.3.8. Power Supply/ Circuit Breaker  
The analyzer operates on 100 VAC, 115 VAC or 230 VAC power at either 50Hz or 60Hz. Individual units are set  
up at the factory to accept any combination of these five attributes. As illustrated in Figure 10-15, power enters  
the analyzer through a standard IEC 320 power receptacle located on the rear panel of the instrument. From  
there it is routed through the On/Off switch located in the lower right corner of the Front Panel. A 6.75 Amp  
circuit breaker is built into the ON/OFF Switch.  
AC power is distributed directly to the sample gas pump. The bench and GFC wheel heaters as well as the GFC  
wheel receive AC power via the relay board.  
AC Line power is converted stepped down and converted to DC power by two DC power supplies. One supplies  
+12 VDC, for valves and the IR source, while a second supply provides +5 VDC and ±15 VDC for logic and  
analog circuitry. All DC voltages are distributed via the relay board.  
CAUTION  
Should the AC power circuit breaker trip, investigate and correct the condition causing  
this situation before turning the analyzer back on.  
ON/OFF  
SWITCH  
AC POWER  
Pressure  
Sensors  
ENTRANCE  
Display  
Keypad  
CPU  
PS 1 (+5 VDC; ±15 VDC)  
RELAY  
BOARD  
KEY  
AC POWER  
DC POWER  
Mother  
Board  
PS 2 (+12 VDC)  
Sync/Demod  
IR Source  
Cooling Fan  
Pump  
M/R &  
Segment  
Sensors  
GFC Wheel  
Motor  
Valve Options  
Heaters  
Figure 10-15: Power Distribution Block Diagram  
05232 Rev B3  
183  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
THEORY OF OPERATION  
Model 360E Instruction Manual  
10.4. Interface  
The analyzer has several ways to communicate the outside world, see Figure 10-16. Users can input data and  
receive information directly via the Front panel keypad and display. Direct communication with the CPU is also  
available by way of the analyzer’s RS232 & RS485 I/O ports or an optional Ethernet port. The analyzer can also  
send and receive different kinds of information via its external digital I/O connectors and the three analog outputs  
located on the rear panel.  
COMM A  
Male  
RS–232 ONLY  
RS-232 or RS–485  
COMM B  
CPU  
Female  
Mother  
Board  
Control Inputs:  
ETHERNET  
OPTION  
1 – 6  
Status Outputs:  
1 – 8  
PC/104 BUS  
Analog Outputs  
KEYBOARD  
A1  
A2  
Optional  
4-20 mA  
I2C BUS  
A3  
I2C BUS  
A4ST  
RELAY  
BOARD  
FRONT PANEL DISPLAY  
Figure 10-16: Interface Block Diagram  
10.4.1. Front Panel Interface  
MODE FIELD  
MESSAGE FIELD  
CONCENTRATION FIELD  
STATUS LED’s  
LOCKING SCREW  
FASTENER  
FASTENER  
SAMPLE  
CAL  
CO2 = 400.0  
SETUP  
SAMPLE A  
RANGE = 500.0 PPM  
<TST TST> CAL  
FAULT  
POWER  
GAS FILTER CORRELATION CO2 ANALYZER- MODEL 360E  
KEY DEFINITIONS  
KEYBOARD  
ON / OFF SWITCH  
Figure 10-17: M360E Front Panel Layout  
184  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Model 360E Instruction Manual  
THEORY OF OPERATION  
The most commonly used method for communicating with the M360E Analyzer is via the instrument’s front panel  
which includes a set of three status LEDs, a vacuum florescent display and a keyboard with 8 context sensitive  
keys.  
10.4.1.1. Analyzer Status LED’s  
Three LEDs are used to inform the user of the instruments basic operating status  
Table 10-3: Front Panel Status LED’s  
NAME  
COLOR  
STATE  
Off  
DEFINITION  
SAMPLE  
Green  
Unit is not operating in sample mode, iDAS is disabled.  
On  
Sample Mode active; Front Panel Display being updated, iDAS data being stored.  
Unit is operating in sample mode, front panel display being updated, iDAS hold-off mode is  
ON, iDAS disabled  
Blinking  
Off  
CAL  
Yellow  
Red  
Auto Cal disabled  
On  
Auto Cal enabled  
Blinking  
Off  
Unit is in calibration mode  
FAULT  
CO2 warnings exist  
Warnings exist  
Blinking  
10.4.1.2. Keyboard  
A row of eight keys just below the vacuum florescent display (see Figure 10-17) is the main method by which the  
user interacts with the analyzer. As the software is operated, labels appear on the bottom row of the display  
directly above each active key, defining the function of that key as it is relevant for the operation being  
performed. Pressing a key causes the associated instruction to be performed by the analyzer.  
Note that the keys do not auto-repeat. In circumstances where the same key must be activated for two  
consecutive operations, it must be released and re-pressed.  
10.4.1.3. Display  
The main display of the analyzer is a vacuum florescent display with two lines of 40 text characters each.  
Information is organized in the following manner (see Figure 10-17):  
Mode Field: Displays the name of the analyzer’s current operating mode.  
Message Field: Displays a variety of informational messages such as warning messages, operation data and  
response messages during interactive tasks.  
Concentration Field: Displays the actual concentration of the sample gas currently being measured by the  
analyzer  
Keypad Definition Field: Displays the definitions for the row of keys just below the display. These definitions  
dynamic, context sensitive and software driven.  
05232 Rev B3  
185  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
THEORY OF OPERATION  
Model 360E Instruction Manual  
10.4.1.4. Keyboard/Display Interface Electronics  
I2C to Relay Board  
Key Press  
Detect  
Display Data  
Decoder  
Display  
Controller  
Display Power  
Watchdog  
Keypad  
Decoder  
I2C Interface  
Serial  
Data  
From 5 VDC  
Power Supply  
Optional  
Maintenance  
LED  
Sample LED  
(Green)  
Maint.  
Switch  
2nd Lang.  
Switch  
Cal LED  
(Yellow)  
2 x 40 CHAR. VACUUM  
FLUORESCENT DISPLAY  
Fault LED  
(Red)  
KEYBOARD  
FRONT PANEL  
Beeper  
Figure 10-18: Keyboard and Display Interface Block Diagram  
The keyboard/display interface electronics of the M360E Analyzer watches the status of the eight front panel  
keys, alerts the CPU when keys are depressed, translates data from parallel to serial and back and manages  
communications between the keyboard, the CPU and the front panel display. Except for the Keyboard interrupt  
status bit, all communication between the CPU and the keyboard/display is handled by way of the instrument’s  
I2C buss. The CPU controls the clock signal and determines when the various devices on the bus are allowed to  
talk or required to listen. Data packets are labeled with addresses that identify for which device the information  
is intended.  
KEYPAD DECODER  
Each key on the front panel communicates with a decoder IC via a separate analog line. When a key is  
depressed the decoder chip notices the change of state of the associated signal; latches and holds the state of  
all eight lines (in effect creating an 8-bit data word); alerts the key-depress-detect circuit (a flip-flop IC);  
translates the 8-bit word into serial data and; sends this to the I2C interface chip.  
KEY-DEPRESS-DETECT CIRCUIT  
This circuit flips the state of one of the inputs to the I2C interface chip causing it to send an interrupt signal to the  
CPU  
186  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Model 360E Instruction Manual  
THEORY OF OPERATION  
I2C INTERFACE CHIP  
This IC performs several functions:  
Using a dedicated digital status bit, it sends an interrupt signal alerting the CPU that new data from the  
keyboard is ready to send.  
Upon acknowledgement by the CPU that it has received the new keyboard data, the I2C interface chip  
resets the key-depress-detect flip-flop.  
In response to commands from the CPU, it turns the front panel status LEDs on and off and activates  
the beeper.  
Informs the CPU when the optional maintenance and second language switches have been opened or  
closed (see Chapter 5 for information on these options).  
DISPLAY DATA DECODER  
This decoder the serial translates the data sent by the CPU (in TTY format) into a bitmapped image which is  
sent over a parallel data bus to the display.  
DISPLAY CONTROLLER  
This circuit manages the interactions between the display data decoder and the display itself. It generates a  
clock pulse that keeps the two devices synchronized. It can also, in response to commands from the CPU turn  
off and/or reset the display.  
Additionally, for analyzers with the optional maintenance switch is installed (See Chapter 5), the display  
controller turns on an LED located on the back of the keyboard interface PCA whenever the instrument is placed  
in maintenance mode.  
DISPLAY POWER WATCHDOG  
The Model 360E’s display can begin to show garbled information or lock-up if the DC voltage supplied to it falls  
too low, even momentarily. To alleviate this, a brown-out watchdog circuit monitors the level of the power supply  
and in the event that the voltage level falls below a certain level, turns the display off, then on resetting it  
I2C LINK TO THE RELAY PCA  
While the CPU’s I2C communication with the relay board is also routed through the keyboard/display interface,  
information passed to and from the relay board via this channel is not recognized by, acted upon or affected by  
the circuitry of the keyboard/display interface.  
05232 Rev B3  
187  
Download from Www.Somanuals.com. All Manuals Search And Download.  
THEORY OF OPERATION  
Model 360E Instruction Manual  
10.5. Software Operation  
The Model 360E Gas Filter Correlation Carbon Dioxide Analyzer is at its heart a high performance, 386-based  
microcomputer running MS-DOS. Inside the DOS shell, special software developed by Teledyne Instruments  
interprets user commands via the various interfaces, performs procedures and tasks, stores data in the CPU’s  
various memory devices and calculates the concentration of the sample gas.  
DOS Shell  
API FIRMWARE  
Analyzer Operations  
Calibration Procedures  
Configuration Procedures  
Memory Handling  
IDAS Records  
Calibration Data  
PC/104 BUS  
Autonomic Systems  
Diagnostic Routines  
System Status Data  
ANALYZER  
HARDWARE  
Interface Handling  
Sensor input Data  
Display Messages  
Measurement  
Algorithm  
Keypad  
Analog Output Data  
RS232 & RS485  
External Digital I/O  
PC/104 BUS  
Linearization Table  
Figure 10-19: Basic Software Operation  
10.5.1. Adaptive Filter  
The M360E software processes the CO2 MEAS and CO2 REF signals, after they are digitized by the  
motherboard, through an adaptive filter built into the software. Unlike other analyzers that average the output  
signal over a fixed time period, the M360E averages over a set number of samples, where each sample is 0.2  
seconds. This is technique is known as boxcar averaging. During operation, the software automatically  
switches between two different length filters based on the conditions at hand. Once triggered, the short filter  
remains engaged for a fixed time period to prevent chattering.  
During conditions of constant or nearly constant concentration the software, by default, computes an average of  
the last 750 samples, or approximately 150 seconds. This provides the calculation portion of the software with  
smooth stable readings. If a rapid change in concentration is detected the filter includes, by default, the last 48  
samples, approximately 10 seconds of data, to allow the analyzer to more quickly respond. If necessary, these  
boxcar lengths can be changed between 1 and 1000 samples but with corresponding tradeoffs in rise time and  
signal-to-noise ratio (contact customer service for more information).  
188  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Model 360E Instruction Manual  
THEORY OF OPERATION  
Two conditions must be simultaneously met to switch to the short filter. First the instantaneous concentration  
must exceed the average in the long filter by a fixed amount. Second the instantaneous concentration must  
exceed the average in the long filter by a portion, or percentage, of the average in the long filter.  
10.5.2. Calibration - Slope and Offset  
Calibration of the analyzer is performed exclusively in software.  
During instrument calibration (see Chapter 7) the user enters expected values for zero and span via the front  
panel keypad and commands the instrument to make readings of calibrated sample gases for both levels. The  
readings taken are adjusted, linearized, and compared to the expected values, With this information the software  
computes values for instrument slope and offset and stores these values in memory for use in calculating the  
CO2 concentration of the sample gas.  
The instrument slope and offset values recorded during the last calibration can be viewed by pressing the  
following keystroke sequence:  
SAMPLE  
RANGE = 50.0 MGM  
CO =XX.XX  
SETUP  
< TST TST > CAL  
SAMPLE  
OFFSET = 0.000  
CO =XX.XX  
SETUP  
< TST TST > CAL  
SAMPLE  
TIME = 16:23:34  
CO =XX.XX  
SETUP  
< TST TST > CAL  
SAMPLE  
SLOPE = 1.000  
CO =XX.XX  
SETUP  
< TST TST > CAL  
10.5.3. Measurement Algorithm  
Once the IR photo-detector is signal is demodulated into CO2 MEAS and CO2 REF by the sync/demod board  
and converted to digital data by the motherboard the M360E analytical software calculates the ratio between  
CO2 MEAS and CO2 REF. This value is compared to a look-up table is used, with interpolation, to linearize the  
response of the instrument. The linearized concentration value is combined with calibration slope and offset  
values, then normalized for changes in sample gas pressure to produce the final CO2 concentration. This is the  
value that is displayed on the instrument front panel display and is stored in memory by the analyzer’s iDAS  
system.  
10.5.4. Temperature and Pressure Compensation  
Changes in pressure can have a noticeable, effect on the CO2 concentration calculation. To account for this, the  
Model 360E software includes a feature which allows the instrument to compensation of the CO2 calculations  
based on changes in ambient pressure.  
The TPC feature multiplies the analyzer’s CO2 concentration by a factor which is based on the difference  
between the ambient pressure of the sample gas normalized to standard atmospheric pressure. As ambient  
pressure increases, the compensated CO2 concentration is increased.  
05232 Rev B3  
189  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
THEORY OF OPERATION  
Model 360E Instruction Manual  
10.5.5. Internal Data Acquisition System (iDAS)  
The iDAS is designed to implement predictive diagnostics that stores trending data for users to anticipate when  
an instrument will require service. Large amounts of data can be stored in non-volatile memory and retrieved in  
plain text format for further processing with common data analysis programs. The iDAS has a consistent user  
interface in all Teledyne Instruments analyzers. New data parameters and triggering events can be added to the  
instrument as needed.  
Depending on the sampling frequency and the number of data parameters the iDAS can store several months of  
data, which are retained even when the instrument is powered off or a new firmware is installed. The iDAS  
permits users to access the data through the instrument’s front panel or the remote interface. The latter can  
automatically download stored data for further processing. For information on using the iDAS, refer to Sections  
6.12.  
User Notes  
190  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 360E Instruction Manual  
TROUBLESHOOTING & REPAIR PROCEDURES  
11. TROUBLESHOOTING & REPAIR PROCEDURES  
This contains a variety of methods for identifying the source of performance problems with the analyzer. Also  
included in this are procedures that are used in repairing the instrument.  
CAUTION  
The operations outlined in this chapter are to be performed by qualified maintenance  
personnel only.  
CAUTION  
Risk of electrical shock. Disconnect power before performing the following  
operations.  
11.1. General Troubleshooting Hints  
The analyzer has been designed so that problems can be rapidly detected, evaluated and repaired. During  
operation, the analyzer continuously performs self-check diagnostics and provides the ability to monitor the key  
operating parameters of the instrument without disturbing monitoring operations.  
A systematic approach to troubleshooting will generally consist of the following four steps:  
1. Note any WARNING MESSAGES and take corrective action as required.  
2. Examine the values of all TEST functions and compare to factory values. Note any major deviations  
from the factory values and take correction action as required.  
3. Use the internal electronic status LED’s to determine whether the CPU and I2C buses are running, and if  
the sync/demodulator and relay board are operating properly. Verify that the DC power supplies are  
operating properly by checking the voltage test points on the relay board. Please note that the  
analyzer’s DC power wiring is color-coded and these colors match the color of the corresponding test  
points on the relay board.  
4. SUSPECT A LEAK FIRST! Data from Teledyne Instruments’ service department indicates that 50% of  
all problems are eventually traced to leaks in the pneumatic connections and gas lines of the analyzer  
itself, the source of zero air, span gases or sample gas delivery system.  
Check for gas flow problems such as clogged or blocked internal/external gas lines, damaged seals,  
punctured gas lines, a damaged pump diaphragm, etc.  
5. Follow the procedures defined in Section 11.5 for confirming that the analyzer’s basic components are  
working (power supplies, CPU, relay board, sync/demod board, keypad, GFC wheel motor, etc.). See  
Figure 3-3 for general layout of components and sub-assemblies in the analyzer. See the wiring  
Interconnect Drawing and Interconnect List, documents 04216 and 04217.  
05232 Rev B3  
191  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
TROUBLESHOOTING & REPAIR PROCEDURES  
Model 360E Instruction Manual  
11.1.1. Interpreting WARNING Messages  
The most common and/or serious instrument failures will result in a warning message being displayed on the  
front panel. Table 11-1 lists warning messages, along with their meaning and recommended corrective action.  
It should be noted that if more than two or three warning messages occur at the same time, it is often an  
indication that some fundamental analyzer sub-system (power supply, relay board, motherboard) has failed  
rather than indication of the specific failures referenced by the warnings. In this case, it is recommended that  
proper operation of power supplies (See Section 11.5.2), the relay board (See Section 11.5.5), and the A/D  
Board (See Section11.4.7.1) be confirmed before addressing the specific warning messages.  
The analyzer will alert the user that a Warning Message is active by displaying the keypad label MSG on the  
Front Panel. In this case the Front panel display will look something like the following:  
SAMPLE  
BENCH TEMP WARNING  
CO2 = XXX.0  
TEST CAL  
MSG CLR SETUP  
The analyzer will also alert the user via the Serial I/O COM port(s) and cause the FAULT LED on the front panel  
to blink.  
To view or clear the various warning messages press:  
SAMPLE  
WHEEL TEMP WARNING  
CAL MSG  
CO2 = XX.XX  
CLR SETUP  
TEST deactivates Warning  
Messages until New warning(s)  
are activated  
TEST  
MSG activates Warning  
SAMPLE  
RANGE=500.00 PPM  
MSG  
CO2 = XX.XX  
CLR SETUP  
Messages.  
<TST TST> keys replaced with  
< TST TST > CAL  
TEST key  
SAMPLE  
WHEEL TEMP WARNING  
CO2 = XX.XX  
Press CLR to clear the  
message currently being  
Displayed.  
< TST TST > CAL  
MSG  
CLR SETUP  
If more than one warning is  
active the next message will  
take its place  
Once the last warning has been  
cleared, the analyzer returns to  
SAMPLE Mode  
Make sure warning messages  
are not due to  
legitimate problems..  
Figure 11-1: Viewing and Clearing Warning Messages  
192  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Model 360E Instruction Manual  
TROUBLESHOOTING & REPAIR PROCEDURES  
Table 11-1: Warning Messages - Indicated Failures  
WARNING  
MESSAGE  
FAULT CONDITION  
POSSIBLE CAUSES  
Bad bench heater  
Bad bench temperature sensor  
Bad relay controlling the bench heater  
Entire relay board is malfunctioning  
The optical bench temp is  
BENCH TEMP  
WARNING  
controlled at 48 2 °C.  
I2C buss malfunction  
NOTE: Box temperature typically runs ~7oc warmer than  
ambient temperature.  
Poor/blocked ventilation to the analyzer.  
Stopped exhaust-fan  
BOX TEMP  
WARNING  
Box Temp is  
< 5 °C or > 48 °C.  
Ambient temperature outside of specified range  
Measured concentration value is too high or low.  
Concentration slope value to high or too low  
Measured concentration value is too high.  
Concentration offset value to high.  
Failed disk on chip  
CANNOT DYN  
SPAN  
CANNOT DYN  
ZERO  
CONFIG  
INITIALIZED  
Dynamic Span operation failed  
Dynamic Zero operation failed  
Configuration and Calibration data  
reset to original Factory state.  
Concentration alarm 1 is enabled  
and the measured CO2 level is  
the set point.  
User erased data  
CONC ALRM1  
WARNING  
Concentration alarm 2 is enabled  
and the measured CO2 level is ≥  
the set point.  
CONC ALRM2  
WARNING  
Failed disk on chip  
User cleared data  
DATA INITIALIZED  
Data Storage in iDAS was erased  
Warning only appears on serial I/O com port(s)  
Front panel display will be frozen, blank or will not respond.  
Failed keyboard  
The CPU is unable to  
Communicate with the Front Panel  
Display /Keyboard  
FRONT PANEL  
WARN  
I2C buss failure  
Loose connector/wiring  
Failed IR photo-detector  
PHOTO TEMP  
WARNING  
PHT DRIVE is  
>4800 mVDC  
Failed sync/demod board  
IR photo-detector improperly attached to the sample chamber  
Bench temp too high.  
Mother Board not detected on  
power up.  
Warning only appears on serial i/o com port(s)  
Front panel display will be frozen, blank or will not respond.  
Massive failure of mother board  
REAR BOARD NOT  
DET  
I2C buss failure  
RELAY BOARD  
WARN  
The CPU cannot communicate with  
the Relay Board.  
Failed relay board  
Loose connectors/wiring  
Failed sample pump  
Blocked sample inlet/gas line  
Dirty particulate filter  
SAMPLE FLOW  
WARN  
Sample flow rate is < 500 cc/min or  
> 1000 cc/min.  
Leak downstream of critical flow orifice  
Failed flow sensor/circuitry  
If sample pressure is < 10 in-hg:  
o Blocked particulate filter  
o Blocked sample inlet/gas line  
o Failed pressure sensor/circuitry  
SAMPLE PRES  
WARN  
Sample Pressure is <10 in-Hg or  
> 35 in-Hg  
Normally 29.92 in-Hg at sea level  
decreasing at 1 in-Hg per 1000 ft of  
altitude (with no flow – pump  
disconnected).  
If sample pressure is > 35 in-hg:  
o Pressurized sample gas. Install vent  
o Blocked vent line on pressurized sample/zero/span gas  
supply  
o Bad pressure sensor/circuitry  
05232 Rev B3  
193  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
TROUBLESHOOTING & REPAIR PROCEDURES  
Model 360E Instruction Manual  
Table 11-1: Warning Messages – Indicated Failures (cont.)  
WARNING  
MESSAGE  
FAULT CONDITION  
POSSIBLE CAUSES  
SAMPLE TEMP  
WARN  
Sample temperature is < 10oC or >  
100oC.  
Ambient temperature outside of specified range  
Failed bench heater  
Failed bench temperature sensor  
Relay controlling the bench heater  
Failed relay board  
I2C buss  
SOURCE  
WARNING  
Occurs when CO2 Ref is <1250  
mVDC or >4950 mVDC.  
GFC wheel stopped  
Failed sync/demod board  
If status LED’s on the sync/demod board ARE flashing the  
cause is most likely a failed:  
IR source  
Either of these conditions will result  
in an invalid M/R ratio.  
Relay board  
I2C buss  
IR photo-detector  
SYSTEM RESET  
The computer has rebooted.  
This message occurs at power on. If you have not cycled the  
power on your instrument:  
o Failed +5 VDC power,  
o Fatal error caused software to restart  
o Loose connector/wiring  
WHEEL TEMP  
WARNING  
The filter wheel temperature is  
Blocked cooling vents below GFC Assembly. Make sure that  
adequate clear space beneath the analyzer.  
Analyzer’s top cover removed  
Wheel heater  
controlled at 68 2 °C  
Wheel temperature sensor  
Relay controlling the wheel heater  
Entire relay board  
I2C buss  
11.1.2. Fault Diagnosis with TEST Functions  
Besides being useful as predictive diagnostic tools, the test functions viewable from the front panel can be used  
to isolate and identify many operational problems when combined with a thorough understanding of the  
analyzer’s theory of operation (see Chapter 10).  
The acceptable ranges for these test functions are listed in the “Nominal Range” column of the analyzer Final  
Test and Validation Data Sheet (p/n 04307) shipped with the instrument. Values outside these acceptable  
ranges indicate a failure of one or more of the analyzer’s subsystems. Functions whose values are still within  
the acceptable range but have significantly changed from the measurement recorded on the factory data sheet  
may also indicate a failure. A worksheet has been provided in Appendix C to assist in recording the value of  
these test functions.  
Table 11-2 contains some of the more common causes for these values to be out of range.  
194  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Model 360E Instruction Manual  
Table 11-2: Test Functions - Indicated Failures  
INDICATED FAILURE(S)  
TROUBLESHOOTING & REPAIR PROCEDURES  
TEST  
FUNCTIONS  
Time of day clock is too fast or slow: To adjust See Section 6.6.  
Battery in clock chip on CPU board may be dead.  
TIME  
Incorrectly configured measurement range(s) could cause response problems with a Data logger or chart  
recorder attached to one of the analog output.  
If the Range selected is too small, the recording device will over range.  
If the Range is too big, the device will show minimal or no apparent change in readings.  
Indicates noise level of instrument or CO2 concentration of sample gas (See Section 11.4.2 for causes).  
RANGE  
STABIL  
If the value displayed is too high the IR Source has become brighter. Adjust the variable gain  
potentiometer on the sync/demod board  
If the value displayed is too low or constantly changing and the CO2 REF is OK:  
o Failed multiplexer on the mother board  
o Failed sync/demod board  
o Loose connector or wiring on sync/demod board  
Flow of purge gas to the GFC wheel housing may have stopped  
If the value displayed is too low or constantly changing and the CO2 REF is BAD:  
o GFC wheel stopped or rotation is too slow  
o Failed sync/demod board IR source  
CO2 MEAS  
&
CO2 REF  
o Failed IR source  
o Failed relay board  
o Failed I2C buss  
o Failed IR photo-detector  
When the analyzer is sampling zero air and the ratio is too low:  
o The reference cell of the GFC wheel is contaminated or leaking.  
o The alignment between the GFC wheel and the segment sensor, the M/R sensor or both is  
incorrect.  
MR RATIO  
o Failed sync/demod board  
o Flow of purge gas to the GFC wheel housing may have stopped  
When the analyzer is sampling zero air and the ratio is too high:  
o Zero air is contaminated  
o Failed IR photo-detector  
See Table 11-1 for SAMPLE PRES WARN  
PRES  
Check for gas flow problems. see Section 11.1.6  
SAMPLE FL  
SAMPLE TEMP should be close to BENCH TEMP. Temperatures outside of the specified range or  
oscillating temperatures are cause for concern  
SAMPLE  
TEMP  
Bench temp control improves instrument noise, stability and drift. Temperatures outside of the specified  
range or oscillating temperatures are cause for concern. See Table 11-1 for BENCH TEMP WARNING  
Wheel temp control improves instrument noise, stability and drift. Outside of set point or oscillating  
temperatures are cause for concern. See Table 11-1 for WHEEL TEMP WARNING  
If the box temperature is out of range, check fan in the power supply module. Areas to the side and rear  
of instrument should allow adequate ventilation. See Table 11-1 for BOX TEMP WARNING.  
If this drive voltage is out of range it may indicate one of several problems:  
- A poor mechanical connection between the various components in inside the detector housing  
- An electronic failure of the IR Photo-Detector’s built-in cooling circuitry, or;  
- A temperature problem inside the analyzer chassis. In this case other temperature warnings would also  
be active such as BENCH TEMP WARNING or BOX TEMP WARNING.  
Values outside range indicate  
BENCH TEMP  
WHEEL TEMP  
BOX TEMP  
PHT DRIVE  
Contamination of the zero air or span gas supply  
Instrument is miss-calibrated  
Blocked gas flow  
SLOPE  
Contaminated or leaking GFC wheel (either chamber)  
Faulty IR photo-detector  
Faulty sample faulty IR photo-detector pressure sensor (P1) or circuitry  
Invalid M/R ratio (see above)  
Bad/incorrect span gas concentration due.  
Values outside range indicate  
Contamination of the zero air supply  
Contaminated or leaking GFC wheel (either chamber)  
OFFSET  
Faulty IR photo-detector  
05232 Rev B3  
195  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
TROUBLESHOOTING & REPAIR PROCEDURES  
Model 360E Instruction Manual  
11.1.3. Using the Diagnostic Signal I/O Function  
The Signal I/O parameters found under the DIAG Menu (See Section 6.9.2 and Appendix A) combined with a  
thorough understanding of the instrument’s Theory of Operation (found in Chapter 10) are useful for  
troubleshooting in three ways:  
The technician can view the raw, unprocessed signal level of the analyzer’s critical inputs and outputs.  
All of the components and functions that are normally under algorithmic control of the CPU can be  
manually exercised.  
The technician can directly control the signal level of the Analog and Digital Output signals.  
This allows the technician to systematically observe the effect of directly controlling these signals on the  
operation of the analyzer. Below in Figure 11-2 is an example of how to use the signal I/O menu to view the raw  
voltage of an input signal or to control the state of an output voltage or control signal. The specific parameter will  
vary depending on the situation.  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
SETUP X.X  
PRIMARY SETUP MENU  
CFG DAS RNGE PASS CLK MORE  
EXIT  
SETUP X.X SECONDARY SETUP MENU  
COMM VARS DIAG  
EXIT  
SAMPLE  
ENTER SETUP PASS : 818  
8
8
1
ENTR EXIT  
DIAG  
SIGNAL I/O  
PREV NEXT  
DIAG I/O  
ENTR  
EXIT  
0 ) EXT_ZERO_CAL=ON  
PREV NEXT JUMP  
PRNT EXIT  
If parameter is an  
input signal  
If parameter is an output  
signal or control  
DIAG I/O  
28) SAMPLE_PRESSURE=2540 MV  
DIAG I/O  
22) WHEEL_HTR=ON  
PREV NEXT JUMP  
PRNT EXIT  
PREV NEXT JUMP  
ON PRNT EXIT  
Toggles parameter  
ON/OFF  
DIAG I/O  
22 ) WHEEL_HTR=OFF  
PREV NEXT JUMP  
OFF PRNT EXIT  
Exit returns to  
DIAG display & all values  
return to software control  
Figure 11-2: Example of Signal I/O Function  
196  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Model 360E Instruction Manual  
TROUBLESHOOTING & REPAIR PROCEDURES  
11.1.4. Internal Electronic Status LED’s  
Several LED’s are located inside the instrument to assist in determining if the analyzer’s CPU, I2C buss and  
relay board, GFC wheel and the sync/demodulator board are functioning properly.  
11.1.4.1. CPU Status Indicator  
DS5, a red LED, that is located on upper portion of the motherboard, just to the right of the CPU board, flashes  
when the CPU is running the main program loop. After power-up, approximately 30 to 60 seconds, DS5 should  
flash on and off. If characters are written to the front panel display but DS5 does not flash then the program files  
have become corrupted. If after 30 – 60 seconds neither the DS5 is flashing or no characters have been written  
to the front panel display then the CPU is bad and must be replaced.  
Mother Board  
P/N 04069  
CPU Status LED  
Figure 11-3: CPU Status Indicator  
11.1.4.2. Sync Demodulator Status LED’s  
Two LED’s located on the Sync/Demod Board and are there to make it obvious that the GFC Wheel is spinning  
and the synchronization signals are present:  
Table 11-3: Sync/Demod Board Status Failure Indications  
LED  
FUNCTION  
M/R Sensor Status  
(Flashes slowly)  
FAULT STATUS  
INDICATED FAILURE(S)  
GFC Wheel is not turning  
M/R Sensor on Opto-Pickup Board failed  
Sync/Demod Board failed  
LED is stuck  
ON or OFF  
D1  
JP 4 Connector/Wiring faulty  
Failed/Faulty +5 VDC Power Supply (PS1)  
GFC Wheel is not turning  
Segment Sensor on Opto-Pickup Board failed  
Sync/Demod Board failed  
JP 4 Connector/Wiring faulty  
Failed/Faulty +5 VDC Power Supply (PS1)  
Segment Sensor  
Status  
LED is stuck  
ON or OFF  
D2  
(Flashes quickly)  
05232 Rev B3  
197  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
TROUBLESHOOTING & REPAIR PROCEDURES  
Model 360E Instruction Manual  
D1 – M/R Sensor Status  
D2 – Segment Sensor Status  
JP4 Connector to Opto-Pickup  
Board  
Figure 11-4: Sync/Demod Board Status LED Locations  
11.1.4.3. Relay Board Status LED’s  
There are eight LED’s located on the Relay Board. The most important of which is D1, which indicates the  
health of the I2C buss. If D1 is blinking the other faults following LED’s can be used in conjunction with DIAG  
menu signal I/O to identify hardware failures of the relays and switches on the relay (See Section 6.9.2 and  
Appendix D).  
Table 11-4: I2C Status LED Failure Indications  
LED  
FUNCTION  
FAULT STATUS  
INDICATED FAILURE(S)  
Failed/Halted CPU  
I2C buss Health  
(Watchdog  
Circuit)  
Continuously ON  
or  
Continuously OFF  
Faulty Mother Board, Keyboard or Relay Board  
Faulty Connectors/Wiring between Mother Board,  
Keyboard or Relay Board  
D1  
(Red)  
Failed/Faulty +5 VDC Power Supply (PS1)  
DC VOLTAGE TEST  
POINTS  
STATUS LED’s  
RELAY PCA  
PN 04135  
Figure 11-5: Relay Board Status LEDs  
198  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Model 360E Instruction Manual  
TROUBLESHOOTING & REPAIR PROCEDURES  
Table 11-5: Relay Board Status LED Failure Indications  
SIGNAL I/O PARAMETER  
ACTIVATED BY VIEW RESULT  
LED  
FUNCTION  
DIAGNOSTIC TECHNIQUE  
Voltage displayed should change. If not:  
Failed Heater  
WHEEL_TEMP Faulty Temperature Sensor  
Failed AC Relay  
D2  
Yellow  
WHEEL  
HEATER  
WHEEL_HEATER  
Faulty Connectors/Wiring  
Voltage displayed should change. If not:  
Failed Heater  
BENCH_TEMP Faulty Temperature Sensor  
Failed AC Relay  
D3  
Yellow  
BENCH  
HEATER  
BENCH_HEATER  
N/A  
Faulty Connectors/Wiring  
D4  
Yellow  
SPARE  
N/A  
N/A  
Sample/Cal Valve should audibly change states. If  
not:  
SAMPLE/CAL  
GAS VALVE  
OPTION  
Failed Valve  
Failed Relay Drive IC on Relay Board  
Failed Relay Board  
D5  
Green  
CAL_VALVE  
N/A  
Faulty +12 VDC Supply (PS2)  
Faulty Connectors/Wiring  
Zero/Span Valve should audibly change states. If  
not:  
ZERO/SPAN  
GAS VALVE  
OPTION  
Failed Valve  
Failed Relay Drive IC on Relay Board  
Failed Relay Board  
Faulty +12 VDC Supply (PS2)  
Faulty Connectors/Wiring  
Shutoff Valve should audibly change states. If not:  
Failed Valve  
Failed Relay Drive IC on Relay Board  
Failed Relay Board  
Faulty +12 VDC Supply (PS2)  
Faulty Connectors/Wiring  
Voltage displayed should change. If not:  
Failed IR Source  
D6  
Green  
SPAN_VALVE  
SHUTOFF_VALVE  
IR_SOURCE  
N/A  
N/A  
D7  
Green  
SHUTOFF  
VALVE OPTION  
Faulty +12 VDC Supply (PS2)  
D8  
Green  
IR SOURCE  
CO2_MEASURE Failed Relay Board  
Failed IR Photo-Detector  
Failed Sync/Demod Board  
Faulty Connectors/Wiring  
05232 Rev B3  
199  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
TROUBLESHOOTING & REPAIR PROCEDURES  
Model 360E Instruction Manual  
11.2. Gas Flow Problems  
The M360E has two main gas flow path, the sample gas flow path and the GFC purge gas flow path. Both are  
controlled by a critical flow orifice. Only the sample gas path is measured and reported. When the IZS or  
zero/span valve options are installed, there are several subsidiary paths but none of those are displayed on the  
front panel or stored by the iDAS.  
With the O2 sensor option installed, third gas flow controlled with a critical flow orifice is added, but this flow is  
not measured or reported.  
In general, flow problems can be divided into three categories:  
Flow is too high  
Flow is greater than zero, but is too low, and/or unstable  
Flow is zero (no flow)  
When troubleshooting flow problems, it is crucial to confirm the actual flow rate without relying on the analyzer’s  
flow display. The use of an independent, external flow meter to perform a flow check as described in Section  
9.3.4 is essential. If this test shows the flow to be correct, check the pressure sensors as described in Section  
11.5.6.5.  
The flow diagrams found in a variety of locations within this manual depicting the M360E in its standard  
configuration and with options installed can help in trouble-shooting flow problems. For your convenience they  
are colleted here.  
11.2.1. M360E Internal Gas Flow Diagrams  
Figure 11-7: M360E – Basic Internal Gas Flow  
200  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Model 360E Instruction Manual  
TROUBLESHOOTING & REPAIR PROCEDURES  
Figure 11-6: Internal Pneumatic Flow OPT 50– Zero/Span/Shutoff Valves  
Figure 11-8: Internal Pneumatic Flow – Zero/Span OPT 52 & 53  
05232 Rev B3  
201  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
TROUBLESHOOTING & REPAIR PROCEDURES  
Model 360E Instruction Manual  
Figure 11.9: M360E – Internal Pneumatics with O2 Sensor Option 65  
11.2.2. Typical Sample Gas Flow Problems  
11.2.2.1. Flow is Zero  
The unit displays a SAMPLE FLOW warning message on the front panel display or the SAMPLE FLOW test  
function reports a zero or very low flow rate.  
Confirm that the sample pump is operating (turning). If not, use an AC voltmeter to make sure that power is  
being supplied to the pump. If no power is present at the electrical leads of the pump.  
1. If AC power is being supplied to the pump, but it is not turning, replace the pump.  
2. If the pump is operating but the unit reports no gas flow, perform a flow check as described in Section  
9.3.4.  
3. If no independent flow meter is available:  
Disconnect the gas lines from both the sample inlet and the exhaust outlet on the rear panel of the  
instrument.  
Make sure that the unit is in basic SAMPLE Mode.  
Place a finger over an Exhaust outlet on the rear panel of the instrument.  
If gas is flowing through the analyzer, you will feel pulses of air being expelled from the Exhaust  
outlet.  
202  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Model 360E Instruction Manual  
TROUBLESHOOTING & REPAIR PROCEDURES  
4. If gas flows through the instrument when it is disconnected from its sources of zero air, span gas or  
sample gas, the flow problem is most likely not internal to the analyzer. Check to make sure that:  
All calibrators/generators are turned on and working correctly.  
Gas bottles are not empty or low.  
Valves, regulators and gas lines are not clogged or dirty.  
11.2.2.2. Low Flow  
1. Check if the pump diaphragm is in good condition. If not, rebuild the pump (See Section 9.3.2). Check  
the Spare Parts List for information of pump rebuild kits.  
2. Check for leaks as described in Section 9.3.3. Repair the leaking fitting, line or valve and re-check.  
3. Check for the sample filter and the orifice filter for dirt. Replace filters (See Sections 9.3.1 and 11.5.1  
respectively).  
4. Check for partially plugged pneumatic lines, orifices, or valves. Clean or replace them.  
5. If an IZS option is installed in the instrument, press CALZ and CALS. If the flow increases then suspect  
a bad sample/cal valve.  
11.2.2.3. High Flow  
The most common cause of high flow is a leak in the sample flow control assembly or between there and the  
pump. If no leaks or loose connections are found in the fittings or the gas line between the orifice and the pump,  
rebuild/clean the sample flow control assembly as described in Section 11.6.1.  
11.2.2.4. Displayed Flow = “XXXX”  
This warning means that there is inadequate gas flow. There are four conditions that might cause this:  
1. A leak upstream or downstream of the flow sensor  
2. A flow obstruction upstream or downstream of the flow sensor  
3. Bad Flow Sensor Board  
4. Bad pump  
To determine which case it is, view the sample pressure and sample flow functions on the front panel. If the  
sample pressure is reading abnormally low, then the cause is likely a flow obstruction upstream of the flow  
sensor. First, check the sample filter and make sure it is not plugged and then systematically check all the other  
components upstream of the orifice to ensure that they are not obstructed.  
If the sample pressure is reading normal but the sample flow is reading low then it is likely that the pump  
diaphragm is worn or there is an obstruction downstream of the flow sensor.  
11.2.2.5. Actual Flow Does Not Match Displayed Flow  
If the actual flow measured does not match the displayed flow, but is within the limits of 720-880 cc/min, adjust  
the calibration of the flow measurement as described in Section 6.9.8.  
05232 Rev B3  
203  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
TROUBLESHOOTING & REPAIR PROCEDURES  
Model 360E Instruction Manual  
11.2.2.6. Sample Pump  
The sample pump should start immediately after the front panel power switch is turned ON. With the Sample  
Inlet plugged, the test function PRES should read about 10”-Hg for a pump in good condition. Readings above  
10” Hg indicate that the pump needs rebuilding. If the test function SAMP FL is greater than 10 cm3/min there is  
a leak in the pneumatic lines.  
11.2.3. Poor or Stopped Flow of Purge Gas  
If sufficient purge gas is not supplied to the GFC wheel housing, cyclical fluctuations in readings at zero or low  
CO2 concentrations, such as < 100 ppm, may occur. These fluctuations are the result of changes in the CO2  
concentration of the ambient atmosphere throughout the course of the day and night. In isolated areas with  
relatively few people working nearby the ambient CO2 concentration will fall during the day and rise during the  
night as rate of photosynthesis of the plants in the surrounding area decreases and increases. In a lab  
environment with a relatively high human occupancy the ambient CO2 concentration will increase during those  
parts of the day when the highest number of workers are present. If the GFC wheel housing is allowed to fill with  
ambient air, these natural, diurnal fluctuations might be detected by the instrument and cause it’s in its  
calculation of the CO2 concentration of the sample gas to drift.  
Another possible symptom of poor or stopped purge gas flow would be the inability to measure zero  
concentrations accurately at the end of a work day on a system that was calibrated at the beginning of a  
workday. Although this could also be due to local fluctuations in ambient CO2 concentration during the day.  
11.3. Calibration Problems  
11.3.1. Miscalibrated  
There are several symptoms that can be caused by the analyzer being miss-calibrated. This condition is  
indicated by out of range Slopes and Offsets as displayed through the test functions and is frequently caused by  
the following:  
1. BAD SPAN GAS. This can cause a large error in the slope and a small error in the offset. Delivered  
from the factory, the M360E’s slope is within ±15% of nominal. Bad span gas will cause the analyzer to  
be calibrated to the wrong value. If in doubt have the span gas checked by and independent lab.  
2. CONTAMINATED ZERO GAS. Excess H2O can cause a positive or negative offset and will indirectly  
affect the slope.  
3. Dilution calibrator not set up correctly or is malfunctioning. This will also cause the slope, but not the  
zero, to be incorrect. Again the analyzer is being calibrated to the wrong value.  
4. Too many analyzers on the manifold. This can cause either a slope or offset error because ambient gas  
with its pollutants will dilute the zero or span gas.  
204  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Model 360E Instruction Manual  
TROUBLESHOOTING & REPAIR PROCEDURES  
11.3.2. Non-Repeatable Zero and Span  
As stated earlier, leaks both in the M360E and in the external system are a common source of unstable and non-  
repeatable readings.  
5. Check for leaks in the pneumatic systems as described in Section 9.3.3. Don’t forget to consider  
pneumatic components in the gas delivery system outside the M360E. Such as:  
A change in zero air source such as ambient air leaking into zero air line, or;  
A change in the span gas concentration due to zero air or ambient air leaking into the span gas line.  
6. Once the instrument passes a leak check, do a flow check (See Section 9.3.4) to make sure adequate  
sample is being delivered to the sensor assembly.  
7. A failing IR photo-detector may be at fault. Check the CO2 MEAS and CO2 REF test functions via the  
front panel display to make sure the signal levels are in the normal range (See Appendix A) and are  
quiet.  
8. Confirm the sample pressure, wheel temperature, bench temperature, and sample flow readings are  
correct and have steady readings.  
9. Disconnect the exhaust line from the optical bench near the rear of the instrument and plug this line into  
the SAMPLE inlet creating a pneumatic loop. The CO2 concentration (either zero or span) now must be  
constant. If readings become quiet, the problem is in the external pneumatics supplies for sample gas,  
span gas or zero air.  
10. If pressurized span gas is being used with a zero/span valve option, make sure that the venting is  
adequate (See Section 3.1.2 and 5.4)  
11. If it is the zero point that is non-repeatable, and if that non-repeatability seems to only occur at a certain  
time of day, such as when worker occupancy is highest or lowest, make sure the flow of purge gas to  
the GFC wheel housing has not stopped (See Sections 10.2.2 and 11.1.7 for more information).  
11.3.3. Inability to Span – No SPAN Key  
1. Confirm that the carbon dioxide span gas source is accurate; this can be done by switching between two  
span-gas tanks. If the CO2 concentration is different, there is a problem with one of the tanks.  
2. Check for leaks in the pneumatic systems as described in Section 9.3.3.  
3. Make sure that the expected span gas concentration entered into the instrument during calibration is the  
correct span gas concentration and not too different from expected span value. This can be viewed via  
the RNG Menu (See Section 6.7).  
4. Check to make sure that there is no ambient air or zero air leaking into span gas line.  
11.3.4. Inability to Zero – No ZERO Key  
1. Confirm that there is a good source of zero air. Dilute a tank of span gas with the same amount of zero  
air from two different sources. If the CO2 Concentration of the two measurements is different, there is a  
problem with one of the sources of zero air.  
2. Check for leaks in the pneumatic systems as described in Section 9.3.3.  
3. If the analyzer has had zero/span valve options 52 or 53, the CO2 scrubber may need maintenance.  
4. Check to make sure that there is no ambient air leaking into zero air line.  
05232 Rev B3  
205  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
TROUBLESHOOTING & REPAIR PROCEDURES  
Model 360E Instruction Manual  
11.4. Other Performance Problems  
Dynamic problems (i.e. problems which only manifest themselves when the analyzer is monitoring sample gas)  
can be the most difficult and time consuming to isolate and resolve. The following provides an itemized list of  
the most common dynamic problems with recommended troubleshooting checks and corrective actions.  
11.4.1. Temperature Problems  
Individual control loops are used to maintain the set point of the absorption bench, filter wheel, and IR photo-  
detector temperatures. If any of these temperatures are out of range or are poorly controlled, the M360E will  
perform poorly.  
11.4.1.1. Box or Sample Temperature  
Box Temperature  
The box temperature sensor is mounted to the motherboard and cannot be disconnected to check its resistance.  
Rather check the BOX TEMP signal using the SIGNAL I/O function under the DIAG Menu (See Section 11.1.3).  
This parameter will vary with ambient temperature, but at ~30oC (6-7° above room temperature) the signal  
should be ~1450 mV.  
Sample Temperature  
The Sample Temperature should closely track the bench temperature. If it does not, locate the sensor, which is  
located at the midpoint of the optical bench in a brass fitting. Unplug the connector labeled “Sample”, and  
measure the resistance of the thermistor; at room temperature (25°C) it should be ~30K Ohms, at operating  
temperature, 48°C, it should be ~ 12K Ohms  
11.4.1.2. Bench Temperature  
There are three possible failures that could cause the Bench temperature to be incorrect.  
1. The heater mounted to the bottom of the Absorption bench is electrically shorted or open. Check the  
resistance of the two heater elements by measuring between pin 2 and 4 (~76 Ohms), and pin 3 and 4  
(~330 Ohms), of the white five-pin connector just below the sample temperature sensor on the Bench  
(pin 1 is the pointed end).  
2. Assuming that the I2C buss is working and that there is no other failure with the relay board, the solid-  
state relay (K2) on the relay board may have failed. Using the BENCH_HEATER parameter under the  
signal I/O function, as described above, turn on and off K2 (D3 on the relay board should illuminate as  
the heater is turned on). Check the AC voltage present between pin 2 and 4, for a 100 or 115 VAC  
model, and pins 3 and 4, for a 220-240 VAC model.  
WARNING:  
HAZARDOUS VOLTAGES ARE PRESENT DURING THIS TEST  
3. If the relay has failed there should be no change in the voltage across pins 2 and 4 or 3 and 4. NOTE:  
K2 is in a socket for easy replacement.  
206  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Model 360E Instruction Manual  
TROUBLESHOOTING & REPAIR PROCEDURES  
4. If K2 checks out OK, the thermistor temperature sensor located on the optical bench near the front of the  
instrument could be at fault. Unplug the connector labeled “Bench”, and measure the resistance of the  
thermistor. At room temperature it should have approximately 30K Ohms resistance near the 48oC set  
point it should have ~12K ohms.  
11.4.1.3. GFC Wheel Temperature  
Like the bench heater above there are three possible causes for the GFC wheel temperature to have failed.  
1. The wheel heater has failed. Check the resistance between pins 1 and 4 on the white five-pin connector  
just below the sample temperature sensor on the bench (pin 1 is the pointed end). It should be  
approximately 275 ohms.  
2.  
Assuming that the I2C buss is working and that there is no other failure with the relay board, the solid-  
state relay (K1) on the relay board may have failed. Using the WHEEL_HEATER parameter under the signal  
I/O function, as described above, turn on and off K1 (D2 on the relay board should illuminate as the heater is  
turned on). Check the AC voltage present between pin 1 and 4.  
WARNING:  
HAZARDOUS VOLTAGES ARE PRESENT DURING THIS TEST  
3. If the relay has failed there should be no change in the voltage across pins 1 and 4. NOTE: K1 is  
socketed for easy replacement.  
4. If K1 checks out OK, the thermistor temperature sensor located at the front of the filter wheel assembly  
may have failed. Unplug the connector labeled “Wheel”, and measure the resistance of the thermistor.  
The resistance near the 68oC set point is ~5.7k ohms.  
11.4.1.4. IR Photo-Detector TEC Temperature  
If the PHT DRIVE test parameter described above in Table 11-2 is out of range there are two four possible  
causes of failure.  
1. The screws retaining the IR photo detector to the absorption bench have become loose. Carefully  
tighten the screws, hand-tight and note whether, after the analyzer has come up to operating  
temperature, whether the PHT DRIVE voltage has returned to an acceptable level.  
2. The two large transistor-type devices mounted to the side of the Absorption Bench have come loose  
from the bench. Tighten the retaining screws and note whether there is an improvement in the PHT  
DRIVE voltage.  
3. The photo-detector has failed. Contact the factory for instructions.  
4. The sync demodulator circuit board has failed. Contact the factor for instructions.  
05232 Rev B3  
207  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
TROUBLESHOOTING & REPAIR PROCEDURES  
Model 360E Instruction Manual  
11.4.2. Excessive Noise  
Noise is continuously monitored in the TEST functions as the STABIL reading and only becomes meaningful  
after sampling a constant gas concentration for at least 10 minutes. Compare the current STABIL reading with  
that recorded at the time of manufacture (included in the M360E Final Test and Validation Data Sheet-p/n 04271  
shipped with the unit from Teledyne Instruments).  
1. The most common cause of excessive noise is leaks. Leak check and flow check the instrument  
described in Section 9.3.  
2. Detector failure – caused by failure of the hermetic seal or over-temperature due to poor heat sinking of  
the detector can to the optical bench. In addition to increased noise due to poor signal-to-noise ratio,  
another indicator of detector failure is a drop in the signal levels of the CO2 MEASURE signal and CO2  
REFERENCE signal.  
3. Sync/Demod Board failure. There are many delicate, high impedance parts on this board. Check the  
CO2 MEAS and CO2 REF Test Functions via the Front Panel Display.  
4. The detector cooler control circuit can fail for reasons similar to the detector itself failing. Symptoms  
would be a change in MR RATIO Test Function when zero air is being sampled.  
Also check the SIGNAL I/O parameter PHT DRIVE. After warm-up, and at 25oC ambient, if PHT DRIVE  
< 4800 mV, the cooler is working properly. If PHT DRIVE is > 4800 mV there is a malfunction.  
5. The +5 and 15 VDC voltages in the M360E are provided by switching power supplies. Switch mode  
supplies create DC outputs by switching the input AC waveform at high frequencies. As the  
components in the switcher age and degrade, the main problem observed is increased noise on the DC  
outputs. If a noisy switcher power supply is suspected, attach an oscilloscope to the DC output test  
points located on the top right hand edge of the Relay board. Look for short period spikes > 100 mV p-p  
on the DC output.  
208  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Model 360E Instruction Manual  
TROUBLESHOOTING & REPAIR PROCEDURES  
11.5. Subsystem Checkout  
The preceding s of this manual discussed a variety of methods for identifying possible sources of failures or  
performance problems within the analyzer. In most cases this included a list of possible causes. This describes  
how to determine individually determine if a certain component or subsystem is actually the cause of the  
problem being investigated.  
11.5.1. AC Mains Configuration  
The analyzer is correctly configured for the AC mains voltage in use if:  
1. The Sample Pump is running.  
2. The GFC wheel motor is spinning. LED’s D1 & D2 (located on the synch/demod PCA) should be  
flashing.  
3. If incorrect power is suspected, check that the correct voltage and frequency is present at the line input  
on the rear panel.  
If the unit is set for 230 VAC and is plugged into 115VAC, or 100VAC the sample pump will not  
start, and the heaters will not come up to temperature.  
If the unit is set for 115 or 100 VAC and is plugged into a 230 VAC circuit, the circuit breaker  
built into the ON/OFF Switch on the Front Panel will trip to the OFF position immediately after power  
is switched on.  
11.5.2. DC Power Supply  
If you have determined that the analyzer’s AC mains power is working, but the unit is still not operating properly,  
there may be a problem with one of the instrument’s switching power supplies. The supplies can have two  
faults, namely no DC output, and noisy output.  
To assist tracing DC Power Supply problems, the wiring used to connect the various printed circuit assemblies  
and DC Powered components and the associated test points on the relay board follow a standard color-coding  
scheme as defined in the following table.  
Table 11-6: DC Power Test Point and Wiring Color Codes  
NAME  
Dgnd  
+5V  
TEST POINT#  
TP AND WIRE COLOR  
1
2
3
4
5
6
7
Black  
Red  
Agnd  
+15V  
-15V  
Green  
Blue  
Yellow  
Purple  
Orange  
+12V  
+12R  
05232 Rev B3  
209  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
TROUBLESHOOTING & REPAIR PROCEDURES  
Model 360E Instruction Manual  
A voltmeter should be used to verify that the DC voltages are correct per the values in the table below, and an  
oscilloscope, in AC mode, with band limiting turned on, can be used to evaluate if the supplies are producing  
excessive noise (> 100 mV p-p).  
Table 11-7: DC Power Supply Acceptable Levels  
CHECK RELAY BOARD TEST POINTS  
POWER  
SUPPLY  
ASSY  
VOLTAG  
MIN V  
MAX V  
FROM TEST POINT  
TO TEST POINT  
E
NAME  
Dgnd  
#
1
3
3
3
1
6
6
NAME  
+5  
#
2
PS1  
PS1  
PS1  
PS1  
PS1  
PS2  
PS2  
+5  
+15  
4.8  
5.25  
16V  
Agnd  
+15  
4
13.5  
-15  
Agnd  
-15V  
5
-14V  
-0.05  
-0.05  
11.75  
-0.05  
-16V  
0.05  
0.05  
12.5  
0.05  
Agnd  
Chassis  
+12  
Agnd  
Dgnd  
Chassis  
+12V  
Dgnd  
1
Dgnd  
N/A  
7
+12V Ret  
+12V Ret  
Dgnd  
1
11.5.3. I2C Bus  
Operation of the I2C buss can be verified by observing the behavior of D1 on the Relay Board in conjunction with  
the performance of the front panel display. Assuming that the DC power supplies are operating properly and the  
wiring from the motherboard to the Keyboard, and the wiring from the keyboard to the Relay board, is intact, the  
I2C buss is operating properly if:  
D1 on the relay board is flashing, or;  
D1 is not flashing but pressing a key on the front panel results in a change to the display.  
11.5.4. Keyboard/Display Interface  
The front panel keyboard, display and Keyboard Display Interface PCA (03975 or 04258) can be verified by  
observing the operation of the display when power is applied to the instrument and when a key is pressed on the  
front panel. Assuming that there are no wiring problems and that the DC power supplies are operating properly:  
1. The vacuum fluorescent display is good if on power-up a “-“ character is visible on the upper left hand  
corner of the display.  
2. The CPU Status LED, DS5, is flashing, See Section 11.1.4.1.  
3. If there is a “-“ character on the display at power-up and D1 on the relay board is flashing then the  
keyboard/display interface PCA is bad.  
4. If the analyzer starts operation with a normal display but pressing a key on the front panel does not  
change the display, then there are three possible problems:  
One or more of the keys is bad,  
The interrupt signal between the Keyboard Display interface and the motherboard is broken, or  
The Keyboard Display Interface PCA is bad.  
210  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Model 360E Instruction Manual  
TROUBLESHOOTING & REPAIR PROCEDURES  
11.5.5. Relay Board  
The relay board PCA (04135) can be most easily checked by observing the condition of the its status LEDs on  
the relay board, as described in Section 11.1.4.3, and the associated output when toggled on and off through  
signal I/O function in the diagnostic menu, See Section 11.1.3.  
1. If the front panel display responds to key presses and D1 on the relay board is NOT flashing then either  
the wiring between the Keyboard and the relay board is bad, or the relay board is bad.  
2. If D1 on the relay board is flashing and the status indicator for the output in question (heater power,  
valve drive, etc.) toggles properly using the signal I/O function, then the associated control device on the  
relay board is bad. Several of the control devices are in sockets and can be easily replaced. The table  
below lists the control device associated with a particular function:  
Table 11-8: Relay Board Control Devices  
CONTROL  
DEVICE  
K1  
FUNCTION  
IN SOCKET  
Wheel Heater  
Bench Heater  
Spare AC Control  
IZS Valves  
Yes  
Yes  
Yes  
Yes  
No  
K2  
K3  
U4  
IR Source Drive  
U5  
The IR source drive output can be verified by measuring the voltage at J16 with the IR source disconnected. It  
should be 11.5± 0.5 VDC.  
11.5.6. Sensor Assembly  
11.5.6.1. Sync/Demodulator Assembly  
To verify that the Sync/Demodulator Assembly is working follow the procedure below:  
1. Verify that D1 and D2 are flashing (they flash at different rates, see Table 11-3).  
If not check the opto pickup assembly, Section 11.5.6.2 and the GFC wheel drive, Section 11.5.6.3.  
If the wheel drive and opto pickup are working properly then verify that there is 2.4 ±0.1 VAC and  
2.5 ±0.15 VDC between digital ground and TP 5 on the sync demod board. If not then check the  
wiring between the sync/demod and opto pickup assembly (see interconnect drawing 04216). If  
good then the sync/demod board is bad.  
2. Verify that the IR source is operating, Section 11.5.6.4.  
3. With the analyzer connected to zero air, measure between TP11 (measure) and analog ground, and  
TP12 (reference) and analog ground.  
If they are similar to values recorded on the factory data sheet then there is likely a problem with the  
wiring or the A/D converter.  
If they are not then either the sync demodulator board or the IR-photodetector are bad. See Section  
11.4.1.4 for problems with the IR-photodetector TEC drive.  
05232 Rev B3  
211  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
TROUBLESHOOTING & REPAIR PROCEDURES  
Model 360E Instruction Manual  
11.5.6.2. Opto Pickup Assembly  
Operation of the opto pickup PCA (04088) can be verified with a voltmeter. Measure the AC and DC voltage  
between digital ground on the relay board, or keyboard and TP1 and TP2 on the sync pickup PCA. For a  
working board, with the GFC motor spinning, they should read 2.4 ±0.1 VAC and 2.5 ±0.15 VDC.  
Further confirmation that the pickups and motor are operating properly can be obtained by measuring the  
frequency at TP1 and TP2 using a frequency counter, a digital volt meter with a frequency counter, or an  
oscilloscope per the table below.  
Table 11-9: Opto Pickup Board Nominal Output Frequencies  
NOMINAL MEASURED FREQUENCY  
AC MAINS FREQ.  
50 Hz  
TP1  
25  
TP2  
300  
360  
60 Hz  
30  
11.5.6.3. GFC Wheel Drive  
If the D1 and D2 on the sync demodulator board are not flashing then:  
1. Check for power to the motor by measuring between pins 1 and 3 on the connector feeding the motor.  
For instruments configured for 120 or 220-240VAC there should be approximately 88 VAC for  
instruments configured for 100VAC, it should be the voltage of the AC mains, approximately 100VAC.  
2. Verify that the frequency select jumper, JP4, is properly set on the Relay Board. For 50 Hz operation it  
should be installed. For 60 Hz operation may either be missing or installed in a vertical orientation.  
3. If there is power to the motor and the frequency select jumper is properly set then the motor is likely bad.  
See Section 11.6.2 for instructions on removing and replacing the GFC assembly that the motor is  
bolted to.  
11.5.6.4. IR Source  
The IR source can be checked using the following procedure:  
1. Disconnect the source and check its resistance when cold. When new, the source should have a cold  
resistance of more than 1.5 Ohms but less than 3.5 Ohms. If not, then the source is bad.  
2. With the source disconnected, energize the analyzer and wait for it to start operating. Measure the drive  
Voltage between pins 1 and 2 on the jack that the source is normally connected to; it should be 11.5 ±  
0.25 VDC. If not, then there is a problem with either the wiring, the Relay Board, or the +12V power  
supply.  
3. If the drive voltage is correct in step 2, then remove the source from the heat sink assembly (2 screws  
on top) and connect to its mating connector. Observe the light being emitted from the source. It should  
be centered at the bottom of the U-shaped element. If there is either no emission or a badly centered  
emission then the source is bad.  
11.5.6.5. Pressure/Flow Sensor Assembly  
The pressure/flow sensor PCA, located on the top of the absorption bench, can be checked with a Voltmeter  
using the following procedure which, assumes that the wiring is intact, and that the motherboard and the power  
supplies are operating properly:  
212  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Model 360E Instruction Manual  
TROUBLESHOOTING & REPAIR PROCEDURES  
1. For Pressure related problems:  
Measure the voltage across C1 it should be 5 ± 0.25 VDC. If not then the board is bad.  
Measure the voltage across TP4 and TP1. With the sample pump disabled it should be 4500 mV  
±250 mV. With the pump energized it should be approximately 200 mV less. If not, then S1, the  
pressure transducer is bad, the board is bad, or there is a pneumatic failure preventing the pressure  
transducer from sensing the absorption cell pressure properly.  
For flow related problems:  
Measure the voltage across TP2 and TP1 it should be 10 ±0.25 VDC. If not then the board is bad.  
Measure the voltage across TP3 and TP1. With proper flow (800 sccm at the sample inlet) this  
should be approximately 4.5V (this voltage will vary with altitude). With flow stopped (sample inlet  
blocked) the voltage should be approximately 1V. If the voltage is incorrect, the flow sensor is bad,  
the board is bad or there is a leak upstream of the sensor.  
11.5.7. Motherboard  
11.5.7.1. A/D Functions  
The simplest method to check the operation of the A-to-D converter on the motherboard is to use the Signal I/O  
function under the DIAG menu to check the two A/D reference voltages and input signals that can be easily  
measured with a voltmeter.  
1. Use the Signal I/O function (See Section 11.1.3 and Appendix A) to view the value of REF_4096_MV  
and REF_GND. If both are within 3 mV of nominal (4096 and 0), and are stable, ±0.5 mV then the basic  
A/D is functioning properly. If not then the motherboard is bad.  
2. Choose a parameter in the Signal I/O function such as SAMPLE_PRESSURE, SAMPLE_FLOW,  
CO2_MEASURE or CO2_REFERENCE. Compare these voltages at their origin (see interconnect  
drawing 04215 and interconnect list 04216) with the voltage displayed through the signal I/O function. If  
the wiring is intact but there is a large difference between the measured and displayed voltage (±10 mV)  
then the motherboard is bad.  
11.5.7.2. Analog Outputs: Voltage  
To verify that the analog outputs are working properly, connect a voltmeter to the output in question and perform  
an analog output step test as described in Section 6.9.3.  
For each of the steps, taking into account any offset that may have been programmed into channel (See Section  
6.9.4), the output should be within 1% of the nominal value listed in the table below except for the 0% step,  
which should be within 2 to 3 mV. If one or more of the steps fails to be within this range then it is likely that  
there has been a failure of the either or both of the DACs and their associated circuitry on the motherboard.  
05232 Rev B3  
213  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
TROUBLESHOOTING & REPAIR PROCEDURES  
Model 360E Instruction Manual  
Table 11-10: Analog Output Test Function - Nominal Values Voltage Outputs  
FULL SCALE OUTPUT VOLTAGE  
1V 5V  
NOMINAL OUTPUT VOLTAGE  
100MV  
10V  
STEP  
%
0
1
2
3
4
5
6
0
0
0
1
2
3
4
5
0
2
20  
40  
60  
80  
100  
20 mV  
40 mV  
60 mV  
80 mV  
100 mV  
0.2  
0.4  
0.6  
0.8  
1.0  
4
6
8
10  
11.5.7.3. Analog Outputs: Current Loop  
To verify that the analog outputs with the optional current mode output are working properly, connect a 250 ohm  
resistor across the outputs and use a voltmeter to measure the output as described in Section 6.9.4.2 and then  
perform an analog output step test as described in Section 6.9.3.  
For each step the output should be within 1% of the nominal value listed in the table below.  
Table 11-11: Analog Output Test Function - Nominal Values Current Outputs  
OUTPUT RANGE  
2 -20  
4 -20  
NOMINAL OUTPUT VALUES  
STEP  
%
0
CURRENT  
2 mA  
5.6  
V(250 OHMS)  
CURRENT  
V(250 OHMS)  
1
2
3
4
5
6
0.5V  
1.4  
2.3  
3.2  
4.1  
5
4
1
20  
40  
60  
80  
100  
7.2  
1.8  
2.6  
3.4  
4.2  
5
9.2  
10.4  
13.6  
16.8  
20  
12.8  
16.4  
20  
214  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Model 360E Instruction Manual  
TROUBLESHOOTING & REPAIR PROCEDURES  
11.5.7.4. Status Outputs  
The procedure below can be used to test the Status outputs:  
1. Connect a jumper between the “D“ pin and the “” pin on the status output connector.  
2. Connect a 1000 ohm resistor between the “+” pin and the pin for the status output that is being tested.  
3. Connect a voltmeter between the “” pin and the pin of the output being tested (see table below).  
Under the DIAGSIGNAL I/O menu (See Section 11.1.3), scroll through the inputs and outputs until you get to  
the output in question. Alternately turn on and off the output noting the voltage on the voltmeter, it should vary  
between 0 volts for ON and 5 volts for OFF.  
Table 11-12: Status Outputs Check  
PIN (LEFT TO RIGHT)  
STATUS  
SYSTEM OK  
CONC VALID  
HIGH RANGE  
ZERO CAL  
SPAN CAL  
DIAG MODE  
ALRM1  
1
2
3
4
5
6
7
8
ALRM2  
11.5.7.5. Control Inputs – Remote Zero, Span  
The control input bits can be tested by the following procedure:  
1. Connect a jumper from the +5 pin on the Status connector to the x5V on the Control In connector.  
2. Connect a second jumper from the ‘-‘ pin on the Status connector to the A pin on the Control In  
connector. The instrument should switch from SAMPLE mode to ZERO CAL R mode.  
3. Connect a second jumper from the ‘-‘ pin on the Status connector to the B pin on the Control In  
connector. The instrument should switch from SAMPLE mode to SPAN CAL R mode.  
4. In each case, the M360E should return to SAMPLE mode when the jumper is removed.  
05232 Rev B3  
215  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
TROUBLESHOOTING & REPAIR PROCEDURES  
Model 360E Instruction Manual  
11.5.8. CPU  
There are two major types of failures associated with the CPU board: complete failure and a failure associated  
with the Disk-On Chip on the CPU board. If either of these failures occur, contact the factory.  
1. For complete failures, assuming that the power supplies are operating properly and the wiring is intact,  
the CPU is bad if on powering the instrument:  
The vacuum fluorescent display shows a dash in the upper left hand corner.  
The CPU Status LED, DS5, is not flashing. (See Section 11.1.4.1.)  
There is no activity from the primary RS-232 port on the rear panel even if “? <ret>” is pressed.  
In some rare circumstances this failure may be caused by a bad IC on the motherboard, specifically  
U57 the large, 44 pin device on the lower right hand side of the board. If this is true, removing U57  
from its socket will allow the instrument to startup but the measurements will be incorrect.  
2. If the analyzer stops part way through initialization (there are words on the vacuum fluorescent display)  
then it is likely that the DOC has been corrupted.  
11.5.9. RS-232 Communications  
11.5.9.1. General RS-232 Troubleshooting  
Teledyne Instruments analyzers use the RS-232 communications protocol to allow the instrument to be  
connected to a variety of computer-based equipment. RS-232 has been used for many years and as equipment  
has become more advanced, connections between various types of hardware have become increasingly difficult.  
Generally, every manufacturer observes the signal and timing requirements of the protocol very carefully.  
Problems with RS-232 connections usually center around 4 general areas:  
1. Incorrect cabling and connectors. See Section 6.11.2 for connector and pin-out information.  
2. The BAUD rate and protocol are incorrectly configured. See Section 6.10.7.  
3. If a modem is being used, additional configuration and wiring rules must be observed. See Section  
6.13.2.6  
4. Incorrect setting of the DTE – DCE Switch is set correctly. See Section 6.10.5  
5. Verify that cable (03596) that connects the serial COM ports of the CPU to J12 of the motherboard is  
properly seated  
216  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Model 360E Instruction Manual  
TROUBLESHOOTING & REPAIR PROCEDURES  
11.5.9.2. Troubleshooting Analyzer/Modem or Terminal Operation  
These are the general steps for troubleshooting problems with a modem connected to a Teledyne Instruments  
analyzer.  
1. Check cables for proper connection to the modem, terminal or computer.  
2. Check to make sure the DTE-DCE is in the correct position as described in Section 6.10.5.  
3. Check to make sure the set up command is correct (See Section 6.13.2.7)  
4. Verify that the Ready to Send (RTS) signal is at logic high. The M360E sets pin 7 (RTS) to greater than  
3 volts to enable modem transmission.  
5. Make sure the BAUD rate, word length, and stop bit settings between modem and analyzer match, See  
Section 6.10.7.  
6. Use the RS-232 test function to send “w” characters to the modem, terminal or computer; See Section  
6.10.8.  
7. Get your terminal, modem or computer to transmit data to the analyzer (holding down the space bar is  
one way); the green LED should flicker as the instrument is receiving data.  
8. Make sure that the communications software or terminal emulation software is functioning properly.  
Further help with serial communications is available in a separate manual “RS-232 Programming Notes”  
Teledyne Instruments part number 013500000.  
05232 Rev B3  
217  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
TROUBLESHOOTING & REPAIR PROCEDURES  
Model 360E Instruction Manual  
11.6. Repair Procedures  
This contains procedures that might need to be performed on rare occasions when a major component of the  
analyzer requires repair or replacement.  
11.6.1. Repairing Sample Flow Control Assembly  
The critical flow orifice is housed in the flow control assembly (Teledyne Instruments part number: 001760400)  
located on the top of the optical bench. A sintered filter protects the jewel orifice so it is unusual for the orifice to  
need replacing, but if it does, or the filter needs replacement please use the following procedure (see the Spare  
Parts list in Appendix B for part numbers and kits):  
1. Turn off power to the analyzer.  
2. Locate the assembly attached to the sample pump, see Figure 3–3.  
3. Disconnect the pneumatic connection from the flow assembly and the assembly from the pump.  
4. Remove the fitting and the components as shown in the exploded view in Figure 11.6.  
5. Replace the o-rings (p/n OR000001) and the sintered filter (p/n FL000001).  
6. If replacing the critical flow orifice itself (P/N 00094100), make sure that the side with the colored window  
(usually red) is facing upstream to the flow gas flow.  
7. Apply new Teflon® tape to the male connector threads  
8. Re-assemble in reverse order.  
9. After reconnecting the power and pneumatic lines, flow check the instrument as described in the Section  
9.3.4.  
Pneumatic Connector, Male 1/8”  
(P/N FT_70  
Spring  
(P/N HW_20)  
Sintered Filter  
(P/N FL_01)  
Critical Flow Orifice  
(P/N 00094100)  
Make sure it is placed with the  
jewel down)  
O-Ring  
(P/N OR_01)  
Purge Housing  
(P/N 000850000)  
Figure 11-10: Critical Flow Restrictor Assembly Disassembly  
218  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Model 360E Instruction Manual  
TROUBLESHOOTING & REPAIR PROCEDURES  
11.6.2. Removing/Replacing the GFC Wheel  
When removing or replacing the GFC Wheel it is important to perform the disassembly in the following order to  
avoid damaging the components:  
1. Turn off the analyzer.  
2. Remove the top cover as described in “Getting Started” Section 3.1.  
3. Open the instrument’s hinged front panel.  
4. Locate the GFC wheel/motor assembly (see Figure 3-3).  
5. unplug the following electronic components:  
The GFC wheel housing temperature sensor;  
GFC wheel heater  
GFC wheel motor power supply  
IR source  
6. Unscrew the purge gas line hex nut and remove the 1/8 inch FEP purge gas line.  
10  
7
8
9
Figure 11-11: Opening the GFC Wheel Housing  
7. Remove the two (2) screws holding the opto-pickup printed circuit assembly to the GFC wheel housing.  
8. Carefully remove the opto-pickup printed circuit assembly.  
9. Remove the four (4) screws holding the GFC wheel motor/heat sink assembly to the GFC wheel  
housing.  
10. Carefully remove the GFC wheel motor/heat sink assembly from the GFC wheel housing.  
11. Remove the one (1) screw fastening the GFC wheel/mask assembly to the GFC motor hub.  
05232 Rev B3  
219  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
TROUBLESHOOTING & REPAIR PROCEDURES  
Model 360E Instruction Manual  
11  
12  
Figure 11-12: Removing the GFC Wheel  
12. Remove the GFC wheel/mask assembly.  
13. Follow the previous steps in reverse order to put the GFC wheel/motor assembly back together.  
11.6.3. Disk-On-Chip Replacement Procedure  
Replacing the Disk-on-Chip, may be necessary in certain rare circumstances or to load new instrument software.  
This will cause all of the instrument configuration parameters and iDAS data to be lost. However a backup copy  
of the operating parameters are stored in a second non-volatile memory and will be loaded into the new the  
Disk-on-Chip on power-up. To change the Disk-on-Chip, follow this procedure.  
1. Turn off power to the instrument.  
2. Fold down the rear panel by loosening the thumbscrews on each side  
3. Locate the Disk-on-Chip in the rightmost socket near the right hand side of the CPU assembly. Remove  
the IC by gently prying it up from the socket.  
4. Reinstall the new Disk-on-Chip, making sure the notch in the end of the chip is facing upward.  
5. Close the rear panel and turn on power to the machine.  
User Notes  
220  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Model 360E Instruction Manual  
A Primer on Electro-Static Discharge  
12. A PRIMER ON ELECTRO-STATIC DISCHARGE  
Teledyne Instruments considers the prevention of damage caused by the discharge of static electricity to be  
extremely important part of making sure that your analyzer continues to provide reliable service for a long time.  
This section describes how static electricity occurs, why it is so dangerous to electronic components and  
assemblies as well as how to prevent that damage from occurring.  
12.1. How Static Charges are Created  
Modern electronic devices such as the types used in the various electronic assemblies of your analyzer, are very  
small, require very little power and operate very quickly. Unfortunately, the same characteristics that allow them  
to do these things also make them very susceptible to damage from the discharge of static electricity. Controlling  
electrostatic discharge begins with understanding how electro-static charges occur in the first place.  
Static electricity is the result of something called triboelectric charging which happens whenever the atoms of the  
surface layers of two materials rub against each other. As the atoms of the two surfaces move together and  
separate, some electrons from one surface are retained by the other.  
Materials  
Makes  
Contact  
Materials  
Separate  
+
+
+
+
PROTONS = 3  
ELECTRONS = 2  
PROTONS = 3  
ELECTRONS = 4  
PROTONS = 3  
ELECTRONS = 3  
PROTONS = 3  
ELECTRONS = 3  
NET CHARGE = -1  
NET CHARGE = +1  
NET CHARGE = 0  
NET CHARGE = 0  
Figure 12-1: Triboelectric Charging  
If one of the surfaces is a poor conductor or even a good conductor that is not grounded, the resulting positive or  
negative charge cannot bleed off and becomes trapped in place, or static. The most common example of  
triboelectric charging happens when someone wearing leather or rubber soled shoes walks across a nylon carpet  
or linoleum tiled floor. With each step, electrons change places and the resulting electro-static charge builds up,  
quickly reaching significant levels. Pushing an epoxy printed circuit board across a workbench, using a plastic  
handled screwdriver or even the constant jostling of StyrofoamTM pellets during shipment can also build hefty  
static charges  
Table 12-1: Static Generation Voltages for Typical Activities  
MEANS OF GENERATION  
Walking across nylon carpet  
Walking across vinyl tile  
Worker at bench  
65-90% RH  
1,500V  
250V  
10-25% RH  
35,000V  
12,000V  
6,000V  
100V  
Poly bag picked up from bench  
1,200V  
20,000V  
Moving around in a chair padded  
with urethane foam  
1,500V  
18,000V  
05232 Rev B3  
221  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
A Primer on Electro-Static Discharge  
Model 360E Instruction Manual  
12.2. How Electro-Static Charges Cause Damage  
Damage to components occurs when these static charges come into contact with an electronic device. Current  
flows as the charge moves along the conductive circuitry of the device and the typically very high voltage levels of  
the charge overheat the delicate traces of the integrated circuits, melting them or even vaporizing parts of them.  
When examined by microscope the damage caused by electro-static discharge looks a lot like tiny bomb craters  
littered across the landscape of the component’s circuitry.  
A quick comparison of the values in Table 12-1 with the those shown in the Table 12-2, listing device  
susceptibility levels, shows why Semiconductor Reliability News estimates that approximately 60% of device  
failures are the result of damage due to electro-static discharge.  
Table 12-2: Sensitivity of Electronic Devices to Damage by ESD  
DAMAGE SUSCEPTIBILITY VOLTAGE  
RANGE  
DEVICE  
DAMAGE BEGINS  
OCCURRING AT  
CATASTROPHIC  
DAMAGE AT  
MOSFET  
VMOS  
10  
100  
30  
1800  
100  
NMOS  
60  
GaAsFET  
EPROM  
60  
2000  
100  
100  
140  
150  
190  
200  
300  
300  
300  
500  
500  
500  
JFET  
7000  
500  
SAW  
Op-AMP  
CMOS  
2500  
3000  
2500  
3000  
7000  
500  
Schottky Diodes  
Film Resistors  
This Film Resistors  
ECL  
SCR  
1000  
2500  
Schottky TTL  
Potentially damaging electro-static discharges can occur:  
Any time a charged surface (including the human body) discharges to a device. Even simple contact of a  
finger to the leads of a sensitive device or assembly can allow enough discharge to cause damage. A  
similar discharge can occur from a charged conductive object, such as a metallic tool or fixture.  
When static charges accumulated on a sensitive device discharges from the device to another surface  
such as packaging materials, work surfaces, machine surfaces or other device. In some cases, charged  
device discharges can be the most destructive.  
A typical example of this is the simple act of installing an electronic assembly into the connector or wiring  
harness of the equipment in which it is to function. If the assembly is carrying a static charge, as it is  
connected to ground a discharge will occur.  
Whenever a sensitive device is moved into the field of an existing electro-static field, a charge may be  
induced on the device in effect discharging the field onto the device. If the device is then momentarily  
grounded while within the electrostatic field or removed from the region of the electrostatic field and  
grounded somewhere else, a second discharge will occur as the charge is transferred from the device to  
ground.  
222  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Model 360E Instruction Manual  
A Primer on Electro-Static Discharge  
12.3. Common Myths About ESD Damage  
I didn’t feel a shock so there was no electro-static discharge: The human nervous system isn’t able  
to feel a static discharge of less than 3500 volts. Most devices are damaged by discharge levels much  
lower than that.  
I didn’t touch it so there was no electro-static discharge: Electro-static charges are fields whose lines  
of force can extend several inches or sometimes even feet away from the surface bearing the charge.  
It still works so there was no damage: Sometimes the damaged caused by electro-static discharge can  
completely sever a circuit trace causing the device to fail immediately. More likely, the trace will be only  
partially occluded by the damage causing degraded performance of the device or worse, weakening the  
trace. This weakened circuit may seem to function fine for a short time, but even the very low voltage and  
current levels of the device’s normal operating levels will eat away at the defect over time causing the  
device to fail well before its designed lifetime is reached.  
These latent failures are often the most costly since the failure of the equipment in which the damaged  
device is installed causes down time, lost data, lost productivity, as well as possible failure and damage to  
other pieces of equipment or property.  
Static Charges can’t build up on a conductive surface: There are two errors in this statement.  
Conductive devices can build static charges if they are not grounded. The charge will be equalized  
across the entire device, but without access to earth ground, they are still trapped and can still build to  
high enough levels to cause damage when they are discharged.  
A charge can be induced onto the conductive surface and/or discharge triggered in the presence of a  
charged field such as a large static charge clinging to the surface of a nylon jacket of someone walking up  
to a workbench.  
As long as my analyzer is properly installed, it is safe from damage caused by static discharges:  
It is true that when properly installed the chassis ground of your analyzer is tied to earth ground and its  
electronic components are prevented from building static electric charges themselves. This does not  
prevent discharges from static fields built up on other things, like you and your clothing, from discharging  
through the instrument and damaging it.  
12.4. Basic Principles of Static Control  
It is impossible to stop the creation of instantaneous static electric charges. It is not, however difficult to prevent  
those charges from building to dangerous levels or prevent damage due to electro-static discharge from  
occurring.  
12.4.1. General Rules  
Only handle or work on all electronic assemblies at a properly set up ESD station. Setting up an ESD safe  
workstation need not be complicated. A protective mat properly tied to ground and a wrist strap are all that is  
needed to create a basic anti-ESD workstation (see figure 12-2).  
W ris t S tra p  
P ro te c tiv e M a t  
G ro u n d P o in t  
Figure 12-2: Basic anti-ESD Work Station  
05232 Rev B3  
223  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
A Primer on Electro-Static Discharge  
Model 360E Instruction Manual  
For technicians that work in the field, special lightweight and portable anti-ESD kits are available from most  
suppliers of ESD protection gear. These include everything needed to create a temporary anti-ESD work area  
anywhere.  
Always wear an Anti-ESD wrist strap when working on the electronic assemblies of your analyzer.  
An anti-ESD wrist strap keeps the person wearing it at or near the same potential as other grounded  
objects in the work area and allows static charges to dissipate before they can build to dangerous levels.  
Anti-ESD wrist straps terminated with alligator clips are available for use in work areas where there is no  
available grounded plug.  
Also, anti-ESD wrist straps include a current limiting resistor (usually around one meg-ohm) that protects  
you should you accidentally short yourself to the instrument’s power supply.  
Simply touching a grounded piece of metal is insufficient. While this may temporarily bleed off static  
charges present at the time, once you stop touching the grounded metal new static charges will  
immediately begin to re-build. In some conditions, a charge large enough to damage a component can  
rebuild in just a few seconds.  
Always store sensitive components and assemblies in anti-ESD storage bags or bins: Even when  
you are not working on them, store all devices and assemblies in a closed anti-Static bag or bin. This will  
prevent induced charges from building up on the device or assembly and nearby static fields from  
discharging through it.  
Use metallic anti-ESD bags for storing and shipping ESD sensitive components and assemblies  
rather than pink-poly bags. The famous, “pink-poly” bags are made of a plastic that is impregnated with  
a liquid (similar to liquid laundry detergent) which very slowly sweats onto the surface of the plastic  
creating a slightly conductive layer over the surface of the bag.  
While this layer may equalizes any charges that occur across the whole bag, it does not prevent the build  
up of static charges. If laying on a conductive, grounded surface, these bags will allow charges to bleed  
away but the very charges that build up on the surface of the bag itself can be transferred through the bag  
by induction onto the circuits of your ESD sensitive device. Also, the liquid impregnating the plastic is  
eventually used up after which the bag is as useless for preventing damage from ESD as any ordinary  
plastic bag.  
Anti-Static bags made of plastic impregnated with metal (usually silvery in color) provide all of the charge  
equalizing abilities of the pink-poly bags but also, when properly sealed, create a Faraday cage that  
completely isolates the contents from discharges and the inductive transfer of static charges.  
Storage bins made of plastic impregnated with carbon (usually black in color) are also excellent at  
dissipating static charges and isolating their contents from field effects and discharges.  
Never use ordinary plastic adhesive tape near an ESD sensitive device or to close an anti-ESD  
bag. The act of pulling a piece of standard plastic adhesive tape, such as Scotch® tape, from its roll will  
generate a static charge of several thousand or even tens of thousands of volts on the tape itself and an  
associated field effect that can discharge through or be induced upon items up to a foot away.  
224  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual  
A Primer on Electro-Static Discharge  
12.4.2. Basic anti-ESD Procedures for Analyzer Repair and Maintenance  
12.4.2.1. Working at the Instrument Rack  
When working on the analyzer while it is in the instrument rack and plugged into a properly grounded power  
supply.  
1. Attach your anti-ESD wrist strap to ground before doing anything else.  
Use a wrist strap terminated with an alligator clip and attach it to a bare metal portion of the instrument  
chassis. This will safely connect you to the same ground level to which the instrument and all of its  
components are connected.  
2. Pause for a second or two to allow any static charges to bleed away.  
3. Open the casing of the analyzer and begin work. Up to this point, the closed metal casing of your  
analyzer has isolated the components and assemblies inside from any conducted or induced static  
charges.  
4. If you must remove a component from the instrument, do not lay it down on a non-ESD preventative  
surface where static charges may lie in wait.  
5. Only disconnect your wrist strap after you have finished work and closed the case of the analyzer.  
12.4.2.2. Working at an Anti-ESD Work Bench.  
When working on an instrument of an electronic assembly while it is resting on an anti-ESD work bench:  
1. Plug your anti-ESD wrist strap into the grounded receptacle of the work station before touching any items  
on the work station and while standing at least a foot or so away. This will allow any charges you are  
carrying to bleed away through the ground connection of the workstation and prevent discharges due to  
field effects and induction from occurring.  
2. Pause for a second or two to allow any static charges to bleed away.  
3. Only open any anti-ESD storage bins or bags containing sensitive devices or assemblies after you have  
plugged your wrist strap into the workstation.  
Lay the bag or bin on the workbench surface.  
Before opening the container, wait several seconds for any static charges on the outside surface of the  
container to be bled away by the workstation’s grounded protective mat.  
4. Do not pick up tools that may be carrying static charges while also touching or holding an ESD Sensitive  
Device.  
Only lay tools or ESD-sensitive devices and assemblies on the conductive surface of your workstation.  
Never lay them down on any non-ESD preventative surface.  
5. Place any static sensitive devices or assemblies in anti-static storage bags or bins and close the bag or  
bin before unplugging your wrist strap.  
6. Disconnecting your wrist strap is always the last action taken before leaving the workbench.  
05232 Rev B3  
225  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
A Primer on Electro-Static Discharge  
Model 360E Instruction Manual  
12.4.2.3. Transferring Components from Rack to Bench and Back  
When transferring a sensitive device from an installed Teledyne Instruments analyzer to an Anti-ESD workbench  
or back:  
1. Follow the instructions listed above for working at the instrument rack and workstation.  
2. Never carry the component or assembly without placing it in an anti-ESD bag or bin.  
3. Before using the bag or container allow any surface charges on it to dissipate:  
If you are at the instrument rack, hold the bag in one hand while your wrist strap is connected to a  
ground point.  
If you are at an anti-ESD workbench, lay the container down on the conductive work surface.  
In either case wait several seconds.  
4. Place the item in the container.  
5. Seal the container. If using a bag, fold the end over and fastening it with anti-ESD tape.  
Folding the open end over isolates the component(s) inside from the effects of static fields.  
Leaving the bag open or simply stapling it shut without folding it closed prevents the bag from forming a  
complete protective envelope around the device.  
6. Once you have arrived at your destination, allow any surface charges that may have built up on the bag  
or bin during travel to dissipate:  
Connect your wrist strap to ground.  
If you are at the instrument rack, hold the bag in one hand while your wrist strap is connected to a  
ground point.  
If you are at a anti-ESD work bench, lay the container down on the conductive work surface  
In either case wait several seconds  
7. Open the container.  
12.4.2.4. Opening Shipments from Teledyne Instruments Customer Service.  
Packing materials such as bubble pack and Styrofoam pellets are extremely efficient generators of static electric  
charges. To prevent damage from ESD, Teledyne Instruments ships all electronic components and assemblies in  
properly sealed anti-ESD containers.  
Static charges will build up on the outer surface of the anti-ESD container during shipping as the packing  
materials vibrate and rub against each other. To prevent these static charges from damaging the components or  
assemblies being shipped make sure that you always unpack shipments from Teledyne Instruments Customer  
Service by:  
1. Opening the outer shipping box away from the anti-ESD work area.  
2. Carry the still sealed ant-ESD bag, tube or bin to the anti-ESD work area.  
3. Follow steps 6 and 7 of Section 12.4.2.3 above when opening the anti-ESD container at the work station.  
4. Reserve the anti-ESD container or bag to use when packing electronic components or assemblies to be  
returned to Teledyne Instruments.  
226  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Model 360E Instruction Manual  
A Primer on Electro-Static Discharge  
12.4.2.5. Packing Components for Return to Teledyne Instruments Customer Service.  
Always pack electronic components and assemblies to be sent to Teledyne Instruments Customer Service in anti-  
ESD bins, tubes or bags.  
WARNING  
DO NOT use pink-poly bags.  
NEVER allow any standard plastic packaging materials to touch the electronic  
component/assembly directly  
This includes, but is not limited to, plastic bubble-pack, Styrofoam peanuts,  
open cell foam, closed cell foam, and adhesive tape  
DO NOT use standard adhesive tape as a sealer. Use ONLY anti-ESD tape  
1. Never carry the component or assembly without placing it in an anti-ESD bag or bin.  
2. Before using the bag or container allow any surface charges on it to dissipate:  
If you are at the instrument rack, hold the bag in one hand while your wrist strap is connected to a  
ground point.  
If you are at an anti-ESD workbench, lay the container down on the conductive work surface.  
In either case wait several seconds.  
3. Place the item in the container.  
4. Seal the container. If using a bag, fold the end over and fastening it with anti-ESD tape.  
Folding the open end over isolates the component(s) inside from the effects of static fields.  
Leaving the bag open or simply stapling it shut without folding it closed prevents the bag from forming a  
complete protective envelope around the device.  
NOTE  
If you do not already have an adequate supply of anti-ESD bags or containers available, Teledyne  
Instruments’ Customer Service department will supply them. Follow the instructions listed above for  
working at the instrument rack and workstation.  
User Notes:  
05232 Rev B3  
227  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
A Primer on Electro-Static Discharge  
Model 360E Instruction Manual  
USER NOTES:  
228  
05232 Rev B3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual  
APPENDIX A - Version Specific Software Documentation  
APPENDIX A - Version Specific Software Documentation  
APPENDIX A-1: Model 360E Software Menu Trees  
APPENDIX A-2: Model 360E Setup Variables Available Via Serial I/O  
APPENDIX A-3: Model 360E Warnings and Test Measurements Via Serial I/O  
APPENDIX A-4: Model 360E Signal I/O Definitions  
APPENDIX A-5: Model 360E iDAS Functions  
05233 Rev G.4  
A-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4  
Model 360E Instruction Manual  
APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4  
Table A-1: M360E Setup Variables, Revision G.4  
SETUP VARIABLE  
NUMERIC  
UNITS  
DEFAULT  
VALUE  
VALUE RANGE  
DESCRIPTION  
Low Access Level Setup Variables (818 password)  
DAS_HOLD_OFF  
Minutes  
15  
0.5–20  
Duration of DAS hold off period.  
Number of digits to display to the  
right of the decimal point for  
concentrations on the display.  
CONC_PRECISION  
3
AUTO, 0, 1, 2, 3, 4  
ON enables remote dynamic zero  
calibration; OFF disables it.  
DYN_ZERO  
DYN_SPAN  
OFF  
OFF  
ON, OFF  
ON, OFF  
ON enables remote dynamic span  
calibration; OFF disables it.  
CLOCK_ADJ  
Sec./Day  
0
-60–60  
Time-of-day clock speed adjustment.  
Medium Access Level Setup Variables (929 password)  
ENGL,SECD,  
EXTN  
Selects the language to use for the  
user interface.  
LANGUAGE_SELECT  
MAINT_TIMEOUT  
ENGL 1  
2
Time until automatically switching out  
of software-controlled maintenance  
mode.  
Hours  
0.1–100  
33 MS, 66 MS,  
133 MS, 266 MS,  
533 MS,1 SEC,  
2 SEC  
Conversion time for  
measure/reference detector channel.  
CONV_TIME  
CO_DWELL  
33 MS 1  
Dwell time before taking measure or  
reference sample.  
Seconds  
0.2  
0.1–30  
Number of samples to take in  
measure or reference mode.  
CO_SAMPLE  
FILT_SIZE  
Samples  
Samples  
Samples  
1
1–30  
750  
48  
1–1000  
1–1000  
Moving average filter size.  
Moving average filter size in adaptive  
mode.  
FILT_ASIZE  
Absolute change to trigger adaptive  
filter.  
FILT_DELTA  
FILT_PCT  
PPM  
%
4
10  
90  
ON  
1
1–1000  
1–100  
Percent change to trigger adaptive  
filter.  
Delay before leaving adaptive filter  
mode.  
FILT_DELAY  
Seconds  
0–180  
ON enables adaptive filter; OFF  
disables it.  
FILT_ADAPT  
ON, OFF  
0.1–30  
ON, OFF  
1–500  
Dwell time before taking each  
sample.  
O2_DWELL 5  
Seconds  
ON enables O2 adaptive filter; OFF  
disables it.  
O2_FILT_ADAPT 5  
O2_FILT_SIZE 5  
O2_FILT_ASIZE 5  
O2_FILT_DELTA 5  
O2_FILT_PCT 5  
ON  
60  
10  
2
O2 moving average filter size in  
normal mode.  
Samples  
Samples  
%
O2 moving average filter size in  
adaptive mode.  
1–500  
Absolute change in O2 concentration  
to shorten filter.  
0.1–100  
0.1–100  
Relative change in O2 concentration  
to shorten filter.  
%
2
A-10  
05233 Rev G.4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual  
APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4  
SETUP VARIABLE  
NUMERIC  
UNITS  
DEFAULT  
VALUE  
VALUE RANGE  
DESCRIPTION  
Delay before leaving O2 adaptive  
filter mode.  
O2_FILT_DELAY 5  
O2_DIL_FACTOR 5  
Seconds  
20  
0–300  
Dilution factor for O2. Used only if is  
dilution enabled with  
FACTORY_OPT variable.  
1
0.1–1000  
PPB,PPM,  
UGM,MGM  
Concentration units for user  
interface.  
USER_UNITS  
DIL_FACTOR  
PPM 1  
1
Dilution factor. Used only if is dilution  
enabled with FACTORY_OPT  
variable.  
0.1–1000  
10–600  
Duration of dark cal. First two-thirds  
is stabilization period; final third is  
measure period.  
DARK_CAL_DURATION  
Seconds  
60  
DARK_MEAS_MV  
DARK_REF_MV  
mV  
mV  
0
0
-1000–1000  
-1000–1000  
Dark offset for measure reading.  
Dark offset for reference reading.  
Target concentration during linearity  
adjustment for range 1.  
LIN_TARGET_CONC1  
LIN_NORM_CONC1  
Conc  
PPM  
300  
300  
1–10000  
Target concentration during linearity  
adjustment normalized for T/P for  
range 1.  
0.01–10000  
Measure/reference ratio measured  
LIN_RATIO1  
1
0.01–100  
during linearity adjustment for range  
1.  
LIN_CORRECT1  
1
0.001–999.999  
1–10000  
Linearity correction factor for range 1.  
Target concentration during linearity  
adjustment for range 2.  
LIN_TARGET_CONC2  
Conc  
300  
Target concentration during linearity  
adjustment normalized for T/P for  
range 2.  
LIN_NORM_CONC2  
LIN_RATIO2  
PPM  
300  
1
0.01–10000  
0.01–100  
Measure/reference ratio measured  
during linearity adjustment for range  
2.  
LIN_CORRECT2  
1
0.001–999.999  
1–10000  
Linearity correction factor for range 2.  
Target concentration during linearity  
adjustment.  
LIN_TARGET_CONC  
Conc  
300  
Target concentration during linearity  
adjustment normalized for T/P.  
LIN_NORM_CONC  
PPM  
300  
0.01–10000  
Measure/reference ratio measured  
during linearity adjustment.  
LIN_RATIO  
LIN_CORRECT  
1
1
0.01–100  
0.001–999.999  
ON, OFF  
Linearity correction factor.  
ON enables CO2 compensation; OFF  
disables it.  
CO2_COMP_ENABLE  
OFF  
CO2 concentration to compensate  
for.  
CO2_COMP_CONC  
%
0
0–20  
CO_CONST1  
CO_CONST2  
500  
100–50000  
0–10  
CO calculation constant.  
CO calculation constant.  
1.448  
Electrical test gain factor for measure  
reading.  
ET_MEAS_GAIN  
ET_REF_GAIN  
1
1
0.0001–9.9999  
0.0001–9.9999  
0–5000  
Electrical test gain factor for  
reference reading.  
Target detector reading during  
electrical test.  
ET_TARGET_DET  
ET_TARGET_CONC  
ET_CONC_RANGE  
mV  
4375  
40  
Target concentration during electrical  
test.  
PPM  
Conc.  
1–10000  
D/A concentration range during  
electrical test.  
50  
0.1–50000  
05233 Rev G.4  
A-11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4  
Model 360E Instruction Manual  
SETUP VARIABLE  
NUMERIC  
UNITS  
DEFAULT  
VALUE  
VALUE RANGE  
DESCRIPTION  
Standard temperature for  
temperature compensation.  
STD_TEMP  
ºK  
321  
1–500  
1–50  
Standard pressure for pressure  
compensation.  
STD_PRESS  
"Hg  
28.1  
48  
Optical bench temperature set point  
and warning limits.  
BENCH_SET  
WHEEL_SET  
ºC  
ºC  
ºC  
0–100  
0–100  
30–70  
Warnings:  
43–53  
68  
Wheel temperature set point and  
warning limits.  
Warnings:  
63–73  
50  
O2 sensor cell temperature set point  
and warning limits.  
O2_CELL_SET 5  
Warnings:  
45–55  
Standard O2 cell temperature for  
temperature compensation.  
STD_O2_CELL_TEMP 5  
CO_SPAN1  
ºK  
323  
40  
1–500  
Target CO concentration during span  
calibration of range 1.  
Conc.  
1–10000  
CO_SLOPE1  
CO_OFFSET1  
1
0
0.001–999.999  
-10–10  
CO slope for range 1.  
CO offset for range 1.  
Target CO concentration during span  
calibration of range 2.  
CO_SPAN2  
Conc.  
40  
1–10000  
CO_SLOPE2  
CO_OFFSET2  
1
0
0.001–999.999  
-10–10  
CO slope for range 2.  
CO offset for range 2.  
Target O2 concentration during span  
calibration.  
O2_TARG_SPAN_CONC 5  
%
20.95  
0.1–100  
O2_SLOPE 5  
O2_OFFSET 5  
%
1
0
0.5–2  
-10–10  
O2 slope.  
O2 offset.  
SNGL, DUAL,  
AUTO  
RANGE_MODE  
SNGL 1  
Range control mode.  
CONC_RANGE1  
O2_RANGE 5  
Conc.  
%
50  
100  
0
0.1–50000  
0.1–500  
D/A concentration range 1.  
O2 concentration range.  
RS232_MODE  
BitFlag  
0–65535  
RS-232 COM1 mode flags. Add  
values to combine flags.  
1 = quiet mode  
2 = computer mode  
4 = enable security  
8 = enable hardware handshaking  
16 = enable Hessen protocol 4  
32 = enable multi-drop  
64 = enable modem  
128 = ignore RS-232 line errors  
256 = disable XON / XOFF support  
512 = disable hardware FIFOs  
1024 = enable RS-485 mode  
2048 = even parity, 7 data bits, 1  
stop bit  
4096 = enable command prompt  
A-12  
05233 Rev G.4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual  
APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4  
SETUP VARIABLE  
BAUD_RATE  
NUMERIC  
UNITS  
DEFAULT  
VALUE  
VALUE RANGE  
DESCRIPTION  
300, 1200, 2400,  
4800, 9600,  
19200 1  
RS-232 COM1 baud rate.  
19200, 38400,  
57600, 115200  
Any character in the  
allowed character set.  
Up to 100 characters  
long.  
RS-232 COM1 modem initialization  
string. Sent verbatim plus carriage  
return to modem on power up or  
manually.  
“AT Y0 &D0 &H0  
&I0 S0=2 &B0 &N6  
&M0 E0 Q1 &W0” 1  
MODEM_INIT  
RS-232 COM2 mode flags.  
RS232_MODE2  
BitFlag  
0
0–65535  
(Same settings as RS232_MODE.)  
300, 1200,  
2400, 4800,  
9600, 19200,  
38400, 57600,  
115200  
BAUD_RATE2  
19200 1  
RS-232 COM2 baud rate.  
Any character in the  
allowed character set.  
Up to 100 characters  
long.  
RS-232 COM2 modem initialization  
string. Sent verbatim plus carriage  
return to modem on power up or  
manually.  
“AT Y0 &D0 &H0  
&I0 S0=2 &B0 &N6  
&M0 E0 Q1 &W0” 1  
MODEM_INIT2  
RS232_PASS  
MACHINE_ID  
Password  
ID  
940331  
320  
0–999999  
0–9999  
RS-232 log on password.  
Unique ID number for instrument.  
Any character in the  
allowed character set.  
Up to 100 characters  
long.  
RS-232 interface command prompt.  
Displayed only if enabled with  
RS232_MODE variable.  
COMMAND_PROMPT  
“Cmd> ” 1  
NONE,  
CO MEASURE,  
CO REFERENCE,  
VACUUM PRESSURE,  
SAMPLE PRESSURE,  
SAMPLE FLOW,  
SAMPLE TEMP,  
BENCH TEMP,  
WHEEL TEMP,  
O2 CELL TEMP 5,  
CHASSIS TEMP,  
PHT DRIVE  
TEST_CHAN_ID  
NONE 1  
Diagnostic analog output ID.  
LOW,  
Range to calibrate during contact  
closure or Hessen calibration.  
REMOTE_CAL_MODE  
PASS_ENABLE  
STABIL_GAS  
LOW 1  
OFF  
CO 1  
10  
HIGH  
ON enables passwords; OFF  
disables them.  
ON, OFF  
CO,  
O2 5  
Selects gas for stability  
measurement.  
Stability measurement sampling  
frequency.  
STABIL_FREQ  
Seconds  
Samples  
1–300  
2–40  
Number of samples in concentration  
stability reading.  
STABIL_SAMPLES  
25  
2500  
Photometer temperature warning  
limits. Set point is not used.  
PHOTO_TEMP_SET  
mV  
0–5000  
Warnings:  
250–4750  
05233 Rev G.4  
A-13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4  
Model 360E Instruction Manual  
SETUP VARIABLE  
SAMP_PRESS_SET  
NUMERIC  
UNITS  
DEFAULT  
VALUE  
VALUE RANGE  
DESCRIPTION  
29.92  
Sample pressure warning limits. Set  
point is not used.  
In-Hg  
cc/m  
0–100  
Warnings:  
15–35  
750  
Sample flow warning limits. Set point  
is not used.  
SAMP_FLOW_SET  
Warnings:  
500–1000  
0–5000  
Slope term to correct sample flow  
rate.  
SAMP_FLOW_SLOPE  
VAC_SAMP_RATIO  
1
0.001–100  
0.1–2  
Maximum vacuum pressure / sample  
pressure ratio for valid sample flow  
calculation.  
0.53  
7.5  
Purge pressure warning limits. Set  
point is not used.  
PURGE_PRESS_SET  
SAMP_TEMP_SET  
BOX_SET  
PSIG  
ºC  
0–100  
0–100  
0–100  
Warnings:  
2.5–12.5  
30  
Sample temperature warning limits.  
Set point is not used.  
Warnings:  
10.1–100  
30  
Internal box temperature warning  
limits. Set point is not used.  
ºC  
Warnings:  
5–48  
Optical bench temperature control  
cycle period.  
BENCH_CYCLE  
BENCH_PROP  
Seconds  
1/ºC  
2
5
0.5–30  
0–100  
Optical bench temperature PID  
proportional coefficient. Proportional  
band is the reciprocal of this setting.  
Optical bench temperature PID  
integral coefficient.  
BENCH_INTEG  
BENCH_DERIV  
WHEEL_CYCLE  
0.5  
2
0–100  
0–100  
0.5–30  
Optical bench temperature PID  
derivative coefficient.  
Wheel temperature control cycle  
period.  
Seconds  
2
Wheel temperature PID proportional  
coefficient. Proportional band is the  
reciprocal of this setting.  
WHEEL_PROP  
1/ºC  
1
0–100  
Wheel temperature PID integral  
coefficient.  
WHEEL_INTEG  
WHEEL_DERIV  
0.035  
0–100  
0–100  
0.5–30  
0–10  
Wheel temperature PID derivative  
coefficient.  
2
O2 cell temperature control cycle  
period.  
O2_CELL_CYCLE 5  
O2_CELL_PROP 5  
O2_CELL_INTEG 5  
O2_CELL_DERIV 5  
TPC_ENABLE  
Seconds  
10  
O2 cell PID temperature control  
proportional coefficient.  
1
0.1  
O2 cell PID temperature control  
integral coefficient.  
0–10  
O2 cell PID temperature control  
derivative coefficient.  
0 (disabled)  
ON  
0–10  
ON enables temperature/ pressure  
compensation; OFF disables it.  
OFF, ON  
OFF, ON  
ON enables concentration  
linearization; OFF disables it.  
CONC_LIN_ENABLE  
ON  
Any character in the  
allowed character set.  
Up to 100 characters  
long.  
SERIAL_NUMBER  
“00000000 ” 1  
Unique serial number for instrument.  
A-14  
05233 Rev G.4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual  
APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4  
SETUP VARIABLE  
DISP_INTENSITY  
NUMERIC  
UNITS  
DEFAULT  
VALUE  
VALUE RANGE  
DESCRIPTION  
HIGH,  
MED,  
LOW,  
DIM  
HIGH 1  
Front panel display intensity.  
ON enables automatic reset of the  
I2C bus in the event of  
communication failures; OFF  
disables automatic reset.  
I2C_RESET_ENABLE  
ON  
OFF, ON  
Time-of-day clock format flags.  
Enclose value in double quotes (")  
when setting from the RS-232  
interface.  
“%a” = Abbreviated weekday name.  
“%b” = Abbreviated month name.  
“%d” = Day of month as decimal  
number (01 – 31).  
“%H” = Hour in 24-hour format (00 –  
23).  
“%I” = Hour in 12-hour format (01 –  
12).  
“%j” = Day of year as decimal  
number (001 – 366).  
Any character in the  
allowed character set.  
Up to 100 characters  
long.  
“TIME=%H:%M:%S  
CLOCK_FORMAT  
“%m” = Month as decimal number  
(01 – 12).  
“%M” = Minute as decimal number  
(00 – 59).  
“%p” = A.M./P.M. indicator for 12-  
hour clock.  
“%S” = Second as decimal number  
(00 – 59).  
“%w” = Weekday as decimal number  
(0 – 6; Sunday is 0).  
“%y” = Year without century, as  
decimal number (00 – 99).  
“%Y” = Year with century, as decimal  
number.  
“%%” = Percent sign.  
Concentration alarm trigger  
sensitivity adjustment.  
ALARM_TRIGGER  
REF_SDEV_LIMIT  
Cycles  
mV  
10  
1–100  
Reference detector standard  
deviation must be below this limit to  
switch out of startup mode.  
50  
0.1–500  
Set Point not used  
Reference source warning limits. Set  
point is not used.  
REF_SOURCE_LIMIT  
mV  
1–5000  
Warnings:  
25–4800  
05233 Rev G.4  
A-15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4  
Model 360E Instruction Manual  
SETUP VARIABLE  
NUMERIC  
UNITS  
DEFAULT  
VALUE  
VALUE RANGE  
DESCRIPTION  
Factory option flags. Add values to  
combine flags.  
1 = enable dilution factor  
2 = zero/span valves installed  
4 = conc. alarms routed to relays  
8 = enable linearity adjustment factor  
16 = display units in concentration  
field  
32 = enable software-controlled  
maintenance mode  
FACTORY_OPT  
BitFlag  
0
0–65535  
64 = span valve installed  
128 = enable switch-controlled  
maintenance mode  
256 = compute only offset during  
zero calibration  
512 = 220 V A/C power  
2048 = enable Internet option 3  
4096 = use “old” style numeric data  
entry menus when editing conc. table  
1
2
3
4
5
Enclose value in double quotes (") when setting from the RS-232 interface  
Hessen protocol  
iChip option  
Must power-cycle instrument for these options to take effect  
O2 option  
SETUP VARIABLE  
NUMERIC  
UNITS  
DEFAULT  
VALUE  
VALUE RANGE  
DESCRIPTION  
Low Access Level Setup Variables (818 password)  
DAS_HOLD_OFF  
Minutes  
15  
0.5–20  
Duration of DAS hold off period.  
Number of digits to display to the  
right of the decimal point for  
concentrations on the display.  
CONC_PRECISION  
3
AUTO, 0, 1, 2, 3, 4  
ON enables remote dynamic zero  
calibration; OFF disables it.  
DYN_ZERO  
DYN_SPAN  
OFF  
OFF  
ON, OFF  
ON, OFF  
ON enables remote dynamic span  
calibration; OFF disables it.  
CLOCK_ADJ  
Sec./Day  
0
-60–60  
Time-of-day clock speed adjustment.  
Medium Access Level Setup Variables (929 password)  
ENGL,SECD,  
EXTN  
Selects the language to use for the  
user interface.  
LANGUAGE_SELECT  
MAINT_TIMEOUT  
ENGL 1  
2
Time until automatically switching out  
of software-controlled maintenance  
mode.  
Hours  
0.1–100  
33 MS, 66 MS,  
133 MS, 266 MS,  
533 MS,1 SEC,  
2 SEC  
Conversion time for  
measure/reference detector channel.  
CONV_TIME  
33 MS 1  
A-16  
05233 Rev G.4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual  
APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4  
SETUP VARIABLE  
CO_DWELL  
NUMERIC  
UNITS  
DEFAULT  
VALUE  
VALUE RANGE  
DESCRIPTION  
Dwell time before taking measure or  
reference sample.  
Seconds  
0.2  
0.1–30  
Number of samples to take in  
measure or reference mode.  
CO_SAMPLE  
FILT_SIZE  
Samples  
Samples  
Samples  
1
1–30  
750  
48  
1–1000  
1–1000  
Moving average filter size.  
Moving average filter size in adaptive  
mode.  
FILT_ASIZE  
Absolute change to trigger adaptive  
filter.  
FILT_DELTA  
FILT_PCT  
PPM  
%
4
10  
90  
ON  
1
1–1000  
1–100  
Percent change to trigger adaptive  
filter.  
Delay before leaving adaptive filter  
mode.  
FILT_DELAY  
Seconds  
0–180  
ON enables adaptive filter; OFF  
disables it.  
FILT_ADAPT  
ON, OFF  
0.1–30  
ON, OFF  
1–500  
Dwell time before taking each  
sample.  
O2_DWELL 5  
Seconds  
ON enables O2 adaptive filter; OFF  
disables it.  
O2_FILT_ADAPT 5  
O2_FILT_SIZE 5  
O2_FILT_ASIZE 5  
O2_FILT_DELTA 5  
O2_FILT_PCT 5  
O2_FILT_DELAY 5  
ON  
60  
10  
2
O2 moving average filter size in  
normal mode.  
Samples  
Samples  
%
O2 moving average filter size in  
adaptive mode.  
1–500  
Absolute change in O2 concentration  
to shorten filter.  
0.1–100  
0.1–100  
0–300  
Relative change in O2 concentration  
to shorten filter.  
%
2
Delay before leaving O2 adaptive  
filter mode.  
Seconds  
20  
Dilution factor for O2. Used only if is  
dilution enabled with  
FACTORY_OPT variable.  
O2_DIL_FACTOR 5  
USER_UNITS  
1
PPM 1  
1
0.1–1000  
PPB,PPM,  
UGM,MGM  
Concentration units for user  
interface.  
Dilution factor. Used only if is dilution  
enabled with FACTORY_OPT  
variable.  
DIL_FACTOR  
0.1–1000  
10–600  
Duration of dark cal. First two-thirds  
is stabilization period; final third is  
measure period.  
DARK_CAL_DURATION  
Seconds  
60  
DARK_MEAS_MV  
DARK_REF_MV  
mV  
mV  
0
0
-1000–1000  
-1000–1000  
Dark offset for measure reading.  
Dark offset for reference reading.  
Target concentration during linearity  
adjustment for range 1.  
LIN_TARGET_CONC1  
LIN_NORM_CONC1  
Conc  
PPM  
300  
300  
1–10000  
Target concentration during linearity  
adjustment normalized for T/P for  
range 1.  
0.01–10000  
Measure/reference ratio measured  
LIN_RATIO1  
1
0.01–100  
during linearity adjustment for range  
1.  
LIN_CORRECT1  
1
0.001–999.999  
1–10000  
Linearity correction factor for range 1.  
Target concentration during linearity  
adjustment for range 2.  
LIN_TARGET_CONC2  
Conc  
300  
Target concentration during linearity  
adjustment normalized for T/P for  
range 2.  
LIN_NORM_CONC2  
PPM  
300  
0.01–10000  
05233 Rev G.4  
A-17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4  
Model 360E Instruction Manual  
SETUP VARIABLE  
LIN_RATIO2  
NUMERIC  
UNITS  
DEFAULT  
VALUE  
VALUE RANGE  
DESCRIPTION  
Measure/reference ratio measured  
during linearity adjustment for range  
2.  
1
0.01–100  
LIN_CORRECT2  
1
0.001–999.999  
1–10000  
Linearity correction factor for range 2.  
Target concentration during linearity  
adjustment.  
LIN_TARGET_CONC  
Conc  
300  
Target concentration during linearity  
adjustment normalized for T/P.  
LIN_NORM_CONC  
PPM  
300  
0.01–10000  
Measure/reference ratio measured  
during linearity adjustment.  
LIN_RATIO  
LIN_CORRECT  
1
1
0.01–100  
0.001–999.999  
ON, OFF  
Linearity correction factor.  
ON enables CO2 compensation; OFF  
disables it.  
CO2_COMP_ENABLE  
OFF  
CO2 concentration to compensate  
for.  
CO2_COMP_CONC  
%
0
0–20  
CO_CONST1  
CO_CONST2  
500  
100–50000  
0–10  
CO calculation constant.  
CO calculation constant.  
1.448  
Electrical test gain factor for measure  
reading.  
ET_MEAS_GAIN  
ET_REF_GAIN  
1
1
0.0001–9.9999  
0.0001–9.9999  
0–5000  
Electrical test gain factor for  
reference reading.  
Target detector reading during  
electrical test.  
ET_TARGET_DET  
ET_TARGET_CONC  
ET_CONC_RANGE  
STD_TEMP  
mV  
4375  
40  
Target concentration during electrical  
test.  
PPM  
Conc.  
ºK  
1–10000  
D/A concentration range during  
electrical test.  
50  
0.1–50000  
1–500  
Standard temperature for  
temperature compensation.  
321  
Standard pressure for pressure  
compensation.  
STD_PRESS  
"Hg  
28.1  
48  
1–50  
Optical bench temperature set point  
and warning limits.  
BENCH_SET  
WHEEL_SET  
ºC  
ºC  
ºC  
0–100  
0–100  
30–70  
Warnings:  
43–53  
68  
Wheel temperature set point and  
warning limits.  
Warnings:  
63–73  
50  
O2 sensor cell temperature set point  
and warning limits.  
O2_CELL_SET 5  
Warnings:  
45–55  
Standard O2 cell temperature for  
temperature compensation.  
STD_O2_CELL_TEMP 5  
CO_SPAN1  
ºK  
323  
40  
1–500  
Target CO concentration during span  
calibration of range 1.  
Conc.  
1–10000  
CO_SLOPE1  
CO_OFFSET1  
1
0
0.001–999.999  
-10–10  
CO slope for range 1.  
CO offset for range 1.  
Target CO concentration during span  
calibration of range 2.  
CO_SPAN2  
Conc.  
40  
1–10000  
CO_SLOPE2  
CO_OFFSET2  
1
0
0.001–999.999  
-10–10  
CO slope for range 2.  
CO offset for range 2.  
Target O2 concentration during span  
calibration.  
O2_TARG_SPAN_CONC 5  
%
20.95  
0.1–100  
A-18  
05233 Rev G.4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual  
APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4  
SETUP VARIABLE  
NUMERIC  
UNITS  
DEFAULT  
VALUE  
VALUE RANGE  
DESCRIPTION  
O2_SLOPE 5  
O2_OFFSET 5  
%
1
0
0.5–2  
-10–10  
O2 slope.  
O2 offset.  
SNGL, DUAL,  
AUTO  
RANGE_MODE  
SNGL 1  
Range control mode.  
CONC_RANGE1  
O2_RANGE 5  
Conc.  
%
50  
100  
0
0.1–50000  
0.1–500  
D/A concentration range 1.  
O2 concentration range.  
RS232_MODE  
BitFlag  
0–65535  
RS-232 COM1 mode flags. Add  
values to combine flags.  
1 = quiet mode  
2 = computer mode  
4 = enable security  
8 = enable hardware handshaking  
16 = enable Hessen protocol 4  
32 = enable multi-drop  
64 = enable modem  
128 = ignore RS-232 line errors  
256 = disable XON / XOFF support  
512 = disable hardware FIFOs  
1024 = enable RS-485 mode  
2048 = even parity, 7 data bits, 1  
stop bit  
4096 = enable command prompt  
300, 1200, 2400,  
4800, 9600,  
BAUD_RATE  
19200 1  
RS-232 COM1 baud rate.  
19200, 38400,  
57600, 115200  
Any character in the  
allowed character set.  
Up to 100 characters  
long.  
RS-232 COM1 modem initialization  
string. Sent verbatim plus carriage  
return to modem on power up or  
manually.  
“AT Y0 &D0 &H0  
&I0 S0=2 &B0 &N6  
&M0 E0 Q1 &W0” 1  
MODEM_INIT  
RS-232 COM2 mode flags.  
RS232_MODE2  
BitFlag  
0
0–65535  
(Same settings as RS232_MODE.)  
300, 1200,  
2400, 4800,  
9600, 19200,  
38400, 57600,  
115200  
BAUD_RATE2  
19200 1  
RS-232 COM2 baud rate.  
Any character in the  
allowed character set.  
Up to 100 characters  
long.  
RS-232 COM2 modem initialization  
string. Sent verbatim plus carriage  
return to modem on power up or  
manually.  
“AT Y0 &D0 &H0  
&I0 S0=2 &B0 &N6  
&M0 E0 Q1 &W0” 1  
MODEM_INIT2  
RS232_PASS  
MACHINE_ID  
Password  
ID  
940331  
320  
0–999999  
0–9999  
RS-232 log on password.  
Unique ID number for instrument.  
Any character in the  
allowed character set.  
Up to 100 characters  
long.  
RS-232 interface command prompt.  
Displayed only if enabled with  
RS232_MODE variable.  
COMMAND_PROMPT  
“Cmd> ” 1  
NONE,  
CO MEASURE,  
TEST_CHAN_ID  
NONE 1  
Diagnostic analog output ID.  
CO REFERENCE,  
VACUUM PRESSURE,  
05233 Rev G.4  
A-19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4  
Model 360E Instruction Manual  
SETUP VARIABLE  
NUMERIC  
UNITS  
DEFAULT  
VALUE  
VALUE RANGE  
DESCRIPTION  
SAMPLE PRESSURE,  
SAMPLE FLOW,  
SAMPLE TEMP,  
BENCH TEMP,  
WHEEL TEMP,  
O2 CELL TEMP 5,  
CHASSIS TEMP,  
PHT DRIVE  
LOW,  
Range to calibrate during contact  
closure or Hessen calibration.  
REMOTE_CAL_MODE  
PASS_ENABLE  
STABIL_GAS  
LOW 1  
OFF  
CO 1  
10  
HIGH  
ON enables passwords; OFF  
disables them.  
ON, OFF  
CO,  
O2 5  
Selects gas for stability  
measurement.  
Stability measurement sampling  
frequency.  
STABIL_FREQ  
Seconds  
Samples  
1–300  
2–40  
Number of samples in concentration  
stability reading.  
STABIL_SAMPLES  
25  
2500  
Photometer temperature warning  
limits. Set point is not used.  
PHOTO_TEMP_SET  
SAMP_PRESS_SET  
SAMP_FLOW_SET  
mV  
0–5000  
0–100  
Warnings:  
250–4750  
29.92  
Sample pressure warning limits. Set  
point is not used.  
In-Hg  
cc/m  
Warnings:  
15–35  
750  
Sample flow warning limits. Set point  
is not used.  
Warnings:  
500–1000  
0–5000  
Slope term to correct sample flow  
rate.  
SAMP_FLOW_SLOPE  
VAC_SAMP_RATIO  
1
0.001–100  
0.1–2  
Maximum vacuum pressure / sample  
pressure ratio for valid sample flow  
calculation.  
0.53  
7.5  
Purge pressure warning limits. Set  
point is not used.  
PURGE_PRESS_SET  
SAMP_TEMP_SET  
BOX_SET  
PSIG  
ºC  
0–100  
0–100  
0–100  
Warnings:  
2.5–12.5  
30  
Sample temperature warning limits.  
Set point is not used.  
Warnings:  
10.1–100  
30  
Internal box temperature warning  
limits. Set point is not used.  
ºC  
Warnings:  
5–48  
Optical bench temperature control  
cycle period.  
BENCH_CYCLE  
BENCH_PROP  
BENCH_INTEG  
Seconds  
1/ºC  
2
5
0.5–30  
0–100  
0–100  
Optical bench temperature PID  
proportional coefficient. Proportional  
band is the reciprocal of this setting.  
Optical bench temperature PID  
integral coefficient.  
0.5  
Optical bench temperature PID  
derivative coefficient.  
BENCH_DERIV  
WHEEL_CYCLE  
2
2
0–100  
Seconds  
0.5–30  
Wheel temperature control cycle  
A-20  
05233 Rev G.4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual  
APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4  
SETUP VARIABLE  
NUMERIC  
UNITS  
DEFAULT  
VALUE  
VALUE RANGE  
DESCRIPTION  
period.  
Wheel temperature PID proportional  
coefficient. Proportional band is the  
reciprocal of this setting.  
WHEEL_PROP  
1/ºC  
1
0–100  
Wheel temperature PID integral  
coefficient.  
WHEEL_INTEG  
WHEEL_DERIV  
0.035  
0–100  
0–100  
0.5–30  
0–10  
Wheel temperature PID derivative  
coefficient.  
2
O2 cell temperature control cycle  
period.  
O2_CELL_CYCLE 5  
O2_CELL_PROP 5  
O2_CELL_INTEG 5  
O2_CELL_DERIV 5  
TPC_ENABLE  
Seconds  
10  
O2 cell PID temperature control  
proportional coefficient.  
1
0.1  
O2 cell PID temperature control  
integral coefficient.  
0–10  
O2 cell PID temperature control  
derivative coefficient.  
0 (disabled)  
ON  
0–10  
ON enables temperature/ pressure  
compensation; OFF disables it.  
OFF, ON  
OFF, ON  
ON enables concentration  
linearization; OFF disables it.  
CONC_LIN_ENABLE  
ON  
Any character in the  
allowed character set.  
Up to 100 characters  
long.  
SERIAL_NUMBER  
DISP_INTENSITY  
“00000000 ” 1  
Unique serial number for instrument.  
Front panel display intensity.  
HIGH,  
MED,  
LOW,  
DIM  
HIGH 1  
ON enables automatic reset of the  
I2C bus in the event of  
communication failures; OFF  
disables automatic reset.  
I2C_RESET_ENABLE  
ON  
OFF, ON  
Time-of-day clock format flags.  
Enclose value in double quotes (")  
when setting from the RS-232  
interface.  
“%a” = Abbreviated weekday name.  
“%b” = Abbreviated month name.  
“%d” = Day of month as decimal  
number (01 – 31).  
“%H” = Hour in 24-hour format (00 –  
23).  
“%I” = Hour in 12-hour format (01 –  
12).  
Any character in the  
allowed character set.  
Up to 100 characters  
long.  
“TIME=%H:%M:%S  
CLOCK_FORMAT  
“%j” = Day of year as decimal  
number (001 – 366).  
“%m” = Month as decimal number  
(01 – 12).  
“%M” = Minute as decimal number  
(00 – 59).  
“%p” = A.M./P.M. indicator for 12-  
hour clock.  
“%S” = Second as decimal number  
(00 – 59).  
“%w” = Weekday as decimal number  
(0 – 6; Sunday is 0).  
“%y” = Year without century, as  
05233 Rev G.4  
A-21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4  
Model 360E Instruction Manual  
SETUP VARIABLE  
NUMERIC  
UNITS  
DEFAULT  
VALUE  
VALUE RANGE  
DESCRIPTION  
decimal number (00 – 99).  
“%Y” = Year with century, as decimal  
number.  
“%%” = Percent sign.  
Concentration alarm trigger  
sensitivity adjustment.  
ALARM_TRIGGER  
REF_SDEV_LIMIT  
Cycles  
mV  
10  
1–100  
Reference detector standard  
deviation must be below this limit to  
switch out of startup mode.  
50  
0.1–500  
Set Point not used  
Reference source warning limits. Set  
point is not used.  
REF_SOURCE_LIMIT  
mV  
1–5000  
Warnings:  
25–4800  
Factory option flags. Add values to  
combine flags.  
1 = enable dilution factor  
2 = zero/span valves installed  
4 = conc. alarms routed to relays  
8 = enable linearity adjustment factor  
16 = display units in concentration  
field  
32 = enable software-controlled  
maintenance mode  
FACTORY_OPT  
BitFlag  
0
0–65535  
64 = span valve installed  
128 = enable switch-controlled  
maintenance mode  
256 = compute only offset during  
zero calibration  
512 = 220 V A/C power  
2048 = enable Internet option 3  
4096 = use “old” style numeric data  
entry menus when editing conc. table  
1
2
3
4
5
Enclose value in double quotes (") when setting from the RS-232 interface  
Hessen protocol  
iChip option  
Must power-cycle instrument for these options to take effect  
O2 option  
A-22  
05233 Rev G.4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual  
APPENDIX A-3: Warnings and Test Functions, Revision G.4  
APPENDIX A-3: Warnings and Test Functions, Revision G.4  
Table A-2: M360E Warning Messages, Revision G.4  
NAME  
MESSAGE TEXT  
DESCRIPTION  
WANALOGCAL  
WBENCHTEMP  
ANALOG CAL WARNING  
BENCH TEMP WARNING  
The A/D or at least one D/A channel has not been calibrated.  
Bench temperature outside of warning limits specified by  
BENCH_SET variable.  
WBOXTEMP  
WDYNSPAN  
WDYNZERO  
BOX TEMP WARNING  
CANNOT DYN SPAN  
CANNOT DYN ZERO  
Chassis temperature outside of warning limits specified by  
BOX_SET variable.  
Contact closure span calibration failed while DYN_SPAN was set  
to ON.  
Contact closure zero calibration failed while DYN_ZERO was set  
to ON.  
WCONCALARM1  
WCONCALARM2  
WCONFIGINIT  
WDATAINIT  
CONC ALARM 1 WARN  
CONC ALARM 2 WARN  
CONFIG INITIALIZED  
DATA INITIALIZED  
Concentration limit 1 exceeded.  
Concentration limit 2 exceeded.  
Configuration storage was reset to factory configuration or erased.  
Data storage was erased.  
WFRONTPANEL  
WO2CELLTEMP 1  
FRONT PANEL WARN  
O2 CELL TEMP WARN  
Firmware is unable to communicate with the front panel.  
O2 sensor cell temperature outside of warning limits specified by  
O2_CELL_SET variable.  
WPHOTOTEMP  
PHOTO TEMP WARNING  
Photometer temperature outside of warning limits specified by  
PHOTO_TEMP_SET variable.  
WREARBOARD  
WRELAYBOARD  
WSAMPFLOW  
REAR BOARD NOT DET  
RELAY BOARD WARN  
SAMPLE FLOW WARN  
Rear board was not detected during power up.  
Firmware is unable to communicate with the relay board.  
Sample flow outside of warning limits specified by  
SAMP_FLOW_SET variable.  
WSAMPPRESS  
WSAMPTEMP  
WSOURCE  
SAMPLE PRESS WARN  
SAMPLE TEMP WARN  
SOURCE WARNING  
Sample pressure outside of warning limits specified by  
SAMP_PRESS_SET variable.  
Sample temperature outside of warning limits specified by  
SAMP_TEMP_SET variable.  
Reference reading minus dark offset outside of warning limits  
specified by  
REF_SOURCE_LIMIT variable.  
WSYSRES  
SYSTEM RESET  
Instrument was power-cycled or the CPU was reset.  
WWHEELTEMP  
WHEEL TEMP WARNING  
Wheel temperature outside of warning limits specified by  
WHEEL_SET variable.  
1
O2 option  
05233 Rev G.4  
A-23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
APPENDIX A-3: Warnings and Test Functions, Revision G.4  
Model 360E Instruction Manual  
Table A-3: M360E Test Functions, Revision G.4  
TEST FUNCTION  
NAME1  
MESSAGE TEXT  
RANGE=50.0 PPM 3  
DESCRIPTION  
RANGE  
D/A range in single or auto-range modes.  
RANGE1  
RANGE2  
O2RANGE1  
RANGE1=50.0 PPM 3  
RANGE2=50.0 PPM 3  
O2 RANGE1=100 % 10  
STABIL=0.0 PPM 3  
CO2 STB=0.0 %  
D/A #1 range in dual range mode.  
D/A #2 range in dual range mode.  
O2 sensor range setting.  
STABILITY  
Concentration stability (standard deviation based on setting of  
STABIL_FREQ and STABIL_SAMPLES).  
O2 STB=0.0 % 10  
CO2MEAS  
CO2REF  
MEAS= 4000 mV  
The demodulated, peak IR detector output during the measure  
portion of the CFG Wheel cycle.  
REF= 2000 mV  
The demodulated, peak IR detector output during the reference  
portion of the CFG wheel cycle.  
MRRATIO  
SAMPPRESS  
PURGEPRESS  
SAMPFLOW  
SAMPTEMP  
BENCHTEMP  
WHEELTEMP  
O2CELLTEMP 1  
BOXTEMP  
MR RATIO=1.100  
PRES=29.9 IN-HG-A  
PURGE=7.5 PSIG  
SAMP FL=751 CC/M  
SAMPLE TEMP=26.8 C  
BENCH TEMP=48.1 C  
WHEEL TEMP=68.1 C  
O2 CELL TEMP=50.2 C  
BOX TEMP=26.8 C  
PHT DRIVE=2500.0 MV  
SLOPE  
Measure/reference ratio.  
Sample pressure.  
Purge pressure  
Sample flow rate.  
Sample temperature.  
Bench temperature.  
Wheel temperature.  
O2 sensor cell temperature.  
Internal chassis temperature.  
Photometer temperature.  
CO2 slope, computed during zero/span calibration.  
PHOTOTEMP  
CO2SLOPE  
CO2 SLOPE=1.0001  
CO2OFFSET  
OFFSET  
CO2 offset, computed during zero/span calibration.  
CO2 OFFSET=0.0001  
O2 SLOPE1=0.980  
O2 OFFSET1=1.79 %  
TEST=1751.4 MV  
O2SLOPE 1  
O2OFFSET 1  
TESTCHAN  
O2 slope, computed during zero/span calibration.  
O2 offset, computed during zero/span calibration.  
Value output to TEST_OUTPUT analog output, selected with  
TEST_CHAN_ID variable.  
CLOCKTIME  
TIME=09:52:20  
Current instrument time of day clock.  
1
Only appears if O2 option is installed.  
A-24  
05233 Rev G.4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual APPENDIX A-4: Signal I/O Definitions for 300E Series Analyzers, Revision G.4  
APPENDIX A-4: Signal I/O Definitions for 300E Series Analyzers, Revision G.4  
Table A-4: Signal I/O Definitions for 300E Series Analyzers, Revision G.4  
SIGNAL NAME  
BIT OR CHANNEL  
NUMBER  
DESCRIPTION  
SYNC_OK  
0
1 = sync. OK  
0 = sync. error  
Spare  
1–7  
Internal outputs, U8, J108, pins 18 = bits 07, default I/O address 322 hex  
ELEC_TEST  
DARK_CAL  
0
1 = electrical test on  
0 = off  
1
1 = dark calibration on  
0 = off  
2–5  
6
Spare  
I2C_RESET  
1 = reset I2C peripherals  
0 = normal  
I2C_DRV_RST  
7
0 = hardware reset 8584 chip  
1 = normal  
Control inputs, U11, J1004, pins 1–6 = bits 0–5, default I/O address 321 hex  
EXT_ZERO_CAL  
EXT_SPAN_CAL  
0
1
2
0 = go into zero calibration  
1 = exit zero calibration  
0 = go into span calibration  
1 = exit span calibration  
0 = remote select high range  
1 = default range  
REMOTE_RANGE_HI  
3–5  
6–7  
Spare  
Always 1  
Control inputs, U14, J1006, pins 16 = bits 05, default I/O address 325 hex  
0–5  
6–7  
Spare  
Always 1  
Control outputs, U17, J1008, pins 18 = bits 07, default I/O address 321 hex  
0–7 Spare  
Control outputs, U21, J1008, pins 912 = bits 03, default I/O address 325 hex  
0–3 Spare  
Alarm outputs, U21, J1009, pins 112 = bits 47, default I/O address 325 hex  
ST_SYSTEM_OK2  
4
5
6
7
1 = system OK  
0 = any alarm condition or in diagnostics mode  
1 = conc. limit 1 exceeded  
0 = conc. OK  
ST_CONC_ALARM_1  
2 + 8  
ST_CONC_ALARM_2  
1 = conc. limit 2 exceeded  
0 = conc. OK  
2 + 8  
Spare  
A status outputs, U24, J1017, pins 18 = bits 07, default I/O address 323 hex  
ST_SYSTEM_OK  
ST_CONC_VALID  
ST_HIGH_RANGE  
ST_ZERO_CAL  
ST_SPAN_CAL  
0
1
2
3
4
0 = system OK  
1 = any alarm condition  
0 = conc. valid  
1 = hold off or other conditions  
0 = high auto-range in use  
1 = low auto-range  
0 = in zero calibration  
1 = not in zero  
0 = in span calibration  
05233 Rev G.4  
A-25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
APPENDIX A-4: Signal I/O Definitions for 300E Series Analyzers, Revision G.4  
Model 360E Instruction Manual  
SIGNAL NAME  
BIT OR CHANNEL  
NUMBER  
DESCRIPTION  
1 = not in span  
ST_DIAG_MODE  
5
6
0 = in diagnostic mode  
1 = not in diagnostic mode  
ST_CONC_ALARM_1  
0 = conc. limit 1 exceeded  
1 = conc. OK  
2
ST_CONC_ALARM_2  
7
0 = conc. limit 2 exceeded  
1 = conc. OK  
2
B status outputs, U27, J1018, pins 18 = bits 07, default I/O address 324 hex  
ST_AUTO_REF 2, 3  
0
0 = in auto-reference mode  
1 = not in auto-reference mode  
1–7  
Spare  
Front panel I2C keyboard, default I2C address 4E hex  
MAINT_MODE  
LANG2_SELECT  
SAMPLE_LED  
CAL_LED  
5 (input)  
0 = maintenance mode  
1 = normal mode  
0 = select second language  
1 = select first language (English)  
0 = sample LED on  
1 = off  
6 (input)  
8 (output)  
9 (output)  
10 (output)  
14 (output)  
0 = cal. LED on  
1 = off  
FAULT_LED  
0 = fault LED on  
1 = off  
AUDIBLE_BEEPER  
0 = beeper on (for diagnostic testing only)  
1 = off  
Relay board digital output (PCF8574), default I2C address 44 hex  
RELAY_WATCHDOG  
WHEEL_HTR  
0
Alternate between 0 and 1 at least every 5 seconds to keep relay board  
active  
1
0 = wheel heater on  
1 = off  
BENCH_HTR  
O2_CELL_HEATER 5  
CAL_VALVE  
2
3
4
5
6
6
7
0 = optical bench heater on  
1 = off  
0 = O2 sensor cell heater on  
1 = off  
0 = let cal. gas in  
1 = let sample gas in  
0 = let span gas in  
1 = let zero gas in  
0 = open zero scrubber valve  
1 = close  
SPAN_VALVE  
ZERO_SCRUB_VALV  
E 2  
SHUTOFF_VALVE  
0 = energize shutoff valve  
1 = de-energize  
IR_SOURCE_ON  
0 = IR source on  
1 = off  
Rear board primary MUX analog inputs  
Sample pressure  
SAMPLE_PRESSURE  
0
1
Vacuum pressure  
VACUUM_PRESSURE  
6
PURGE_PRESSURE 9,  
1
Purge pressure  
10  
CO_MEASURE  
2
3
Detector measure reading  
Detector reference reading  
CO_REFERENCE  
A-26  
05233 Rev G.4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual APPENDIX A-4: Signal I/O Definitions for 300E Series Analyzers, Revision G.4  
SIGNAL NAME  
BIT OR CHANNEL  
NUMBER  
DESCRIPTION  
4
5
Temperature MUX  
Sample flow  
SAMPLE_FLOW  
PHOTO_TEMP  
TEST_INPUT_7  
TEST_INPUT_8  
REF_4096_MV  
O2_SENSOR 5  
6
Photometer detector temperature  
Diagnostic test input  
Diagnostic test input  
4.096V reference from MAX6241  
O2 concentration sensor  
Spare  
7
8
9
10  
11  
12  
13  
14  
15  
CO2_SENSOR 7  
CO2 concentration sensor  
Spare  
DAC loopback MUX  
Ground reference  
REF_GND  
Rear board temperature MUX analog inputs  
Internal box temperature  
Sample temperature  
BOX_TEMP  
SAMPLE_TEMP  
BENCH_TEMP  
WHEEL_TEMP  
TEMP_INPUT_4  
TEMP_INPUT_5  
O2_CELL_TEMP  
0
1
2
3
4
5
6
7
Optical bench temperature  
Wheel temperature  
Diagnostic temperature input  
Diagnostic temperature input  
O2 sensor cell temperature  
Spare  
Rear board DAC MUX analog inputs  
DAC channel 0 loopback  
DAC channel 1 loopback  
DAC channel 2 loopback  
DAC channel 3 loopback  
Rear board analog outputs  
Concentration output #1  
Concentration output #2  
Test measurement output  
Concentration output #3 (CO2)  
DAC_CHAN_0  
DAC_CHAN_1  
DAC_CHAN_2  
DAC_CHAN_3  
0
1
2
3
CONC_OUT_1  
CONC_OUT_2  
TEST_OUTPUT  
CONC_OUT_3 7  
Hessen protocol.  
M300EH.  
M300ES.  
M320E.  
O2 option.  
0
1
2
3
1
2
3
4
5
6
Sample pressure or differential pressure flow measurement option.  
7
M306E.  
8
Factory option enables concentration alarms on relay outputs.  
9
M360E.  
GFC7000E.  
10  
05233 Rev G.4  
A-27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
APPENDIX A-5: M360E iDAS Parameters Revision G.4  
Model 360E Instruction Manual  
APPENDIX A-5: M360E iDAS Parameters Revision G.4  
Table A-5: M360E DAS Trigger Events, Revision G.4  
NAME  
ATIMER  
EXITZR  
DESCRIPTION  
Automatic timer expired  
Exit zero calibration mode  
Exit span calibration mode  
Exit multi-point calibration mode  
Slope and offset recalculated  
Exit diagnostic mode  
EXITSP  
EXITMP  
SLPCHG  
EXITDG  
SOURCW  
CONCW1  
CONCW2  
SYNCW  
Source warning  
Concentration limit 1 exceeded  
Concentration limit 2 exceeded  
Sync warning  
BNTMPW  
WTEMPW  
STEMPW  
SFLOWW  
SPRESW  
BTEMPW  
PTEMPW  
Bench temperature warning  
Wheel temperature warning  
Sample temperature warning  
Sample flow warning  
Sample pressure warning  
Box temperature warning  
Photometer detector temperature warning  
A-28  
05233 Rev G.4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual  
APPENDIX A-5: M360E iDAS Parameters Revision G.4  
Table A-6: M360E iDAS Functions, Revision G.4  
NAME  
DETMES  
DETREF  
RATIO  
DESCRIPTION  
UNITS  
mV  
Detector measure reading  
Detector reference reading  
M/R ratio.  
mV  
None  
None  
None  
None  
None  
PPM  
SLOPE1  
SLOPE2  
OFSET1  
OFSET2  
ZSCNC1  
Slope for range #1  
Slope for range #2  
Offset for range #1  
Offset for range #2  
Concentration for range #1 during zero/span calibration, just before computing  
new slope and offset  
ZSCNC2  
Concentration for range #2 during zero/span calibration, just before computing  
new slope and offset  
PPM  
CONC1  
CONC2  
STABIL  
BNTEMP  
WTEMP  
SMPTMP  
SMPFLW  
SMPPRS  
BOXTMP  
PHTDRV  
TEST7  
Concentration for range #1  
PPM  
PPM  
PPM  
°C  
Concentration for range #2  
Concentration stability  
Bench temperature  
Wheel temperature  
°C  
Sample temperature  
°C  
Sample flow  
cc/m  
"Hg  
°C  
Sample pressure  
Internal box temperature  
Photometer detector temperature drive  
Diagnostic test input (TEST_INPUT_7)  
Diagnostic test input (TEST_INPUT_8)  
Diagnostic temperature input (TEMP_INPUT_4)  
Diagnostic temperature input (TEMP_INPUT_5)  
Ground reference (REF_GND)  
4096 mV reference (REF_4096_MV)  
Bench temperature control duty cycle  
mV  
mV  
mV  
°C  
TEST8  
TEMP4  
TEMP5  
°C  
REFGND  
RF4096  
BNCDTY  
mV  
mV  
Fraction  
0 = off,  
1 = on  
WHLDTY  
Wheel temperature control duty cycle  
Fraction  
05233 Rev G.4  
A-29  
Download from Www.Somanuals.com. All Manuals Search And Download.  
APPENDIX A-6: Terminal Command Designators, Revision G.4  
Model 360E Instruction Manual  
APPENDIX A-6: Terminal Command Designators, Revision G.4  
Table A-7: Terminal Command Designators, Revision G.4  
COMMAND  
? [ID]  
ADDITIONAL COMMAND SYNTAX  
DESCRIPTION  
Display help screen and commands list  
Establish connection to instrument  
Terminate connection to instrument  
Display test(s)  
LOGON [ID]  
LOGOFF [ID]  
password  
SET ALL|name|hexmask  
LIST [ALL|name|hexmask] [NAMES|HEX]  
name  
Print test(s) to screen  
T [ID]  
Print single test  
CLEAR ALL|name|hexmask  
SET ALL|name|hexmask  
LIST [ALL|name|hexmask] [NAMES|HEX]  
name  
Disable test(s)  
Display warning(s)  
Print warning(s)  
W [ID]  
Clear single warning  
CLEAR ALL|name|hexmask  
ZERO|LOWSPAN|SPAN [1|2]  
ASEQ number  
Clear warning(s)  
Enter calibration mode  
Execute automatic sequence  
Compute new slope/offset  
Exit calibration mode  
C [ID]  
COMPUTE ZERO|SPAN  
EXIT  
ABORT  
Abort calibration sequence  
Print all I/O signals  
LIST  
name[=value]  
Examine or set I/O signal  
Print names of all diagnostic tests  
Execute diagnostic test  
Exit diagnostic test  
LIST NAMES  
ENTER name  
EXIT  
RESET [DATA] [CONFIG] [exitcode]  
PRINT ["name"] [SCRIPT]  
RECORDS ["name"]  
Reset instrument  
D [ID]  
Print iDAS configuration  
Print number of iDAS records  
REPORT ["name"] [RECORDS=number] [FROM=<start  
date>][TO=<end date>][VERBOSE|COMPACT|HEX]  
(Print DAS records)(date format: MM/DD/YYYY(or YY)  
[HH:MM:SS]  
Print iDAS records  
CANCEL  
Halt printing iDAS records  
Print setup variables  
LIST  
name[=value [warn_low [warn_high]]]  
name="value"  
Modify variable  
Modify enumerated variable  
Print instrument configuration  
Enter/exit maintenance mode  
Print current instrument mode  
Upload iDAS configuration  
Upload single iDAS channel  
Delete iDAS channels  
V [ID]  
CONFIG  
MAINT ON|OFF  
MODE  
DASBEGIN [<data channel definitions>] DASEND  
CHANNELBEGIN propertylist CHANNELEND  
CHANNELDELETE ["name"]  
The command syntax follows the command type, separated by a space character. Strings in [brackets] are optional  
designators. The following key assignments also apply.  
A-30  
05233 Rev G.4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual  
APPENDIX A-6: Terminal Command Designators, Revision G.4  
Table A-8: Terminal Key Assignments, Revision G.4  
TERMINAL KEY ASSIGNMENTS  
ESC  
Abort line  
CR (ENTER)  
Ctrl-C  
Execute command  
Switch to computer mode  
COMPUTER MODE KEY ASSIGNMENTS  
LF (line feed)  
Ctrl-T  
Execute command  
Switch to terminal mode  
USER NOTES:  
05233 Rev G.4  
A-31  
Download from Www.Somanuals.com. All Manuals Search And Download.  
APPENDIX A-6: Terminal Command Designators, Revision G.4  
Model 360E Instruction Manual  
USER NOTES:  
A-32  
05233 Rev G.4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual  
APPENDIX B - M360E Spare Parts List  
APPENDIX B - M360E Spare Parts List  
NOTE  
Use of replacement parts other than those supplied by API may result in non-compliance with  
European standard EN 61010-1.  
05390 - M360E Spare Parts List  
04411 - M360E Recommended Spare Parts Stocking Levels  
05234 Rev B  
B-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
APPENDIX B - M360E Spare Parts List  
Model 360E Instruction Manual  
B-2  
05234 Rev B  
Download from Www.Somanuals.com. All Manuals Search And Download.  
M360E Spare Parts List  
Part Number  
000941000  
Description  
ORIFICE, 13 MIL (SAMPLE FLOW)  
001760400  
001761300  
003290500  
003291500  
006110200  
006900000  
009450300  
009550500  
009560201  
009600400  
009690000  
009690100  
009840300  
010790000  
010800000  
016290000  
016300600  
019340200  
026060000  
026070000  
035280000  
036110300  
037250100  
037600000  
037860000  
040010000  
040030100  
041350000  
041710000  
042410100  
042410200  
042580000  
042680000  
042690000  
042890100  
042900100  
043250100  
043250300  
043250400  
043940000  
045830000  
045840000  
045930000  
048620200  
050110000  
ASSY, FLOW CTL, 800CC, 1/4" CONN-B  
ASSY, SPAN GAS FLOW CONTROL  
ASSY, THERMISTOR, M300 BENCH  
ASSY, THERMISTOR, BENCH/WHEEL, M300E  
ASSY, M300 MOTOR WHEEL HEATER  
PAD, SCRUBBER RETAINER  
ASSY, ZERO/SPAN VALVES, M300E (KB)  
ASSY, SOURCE, M300E S/N>65  
FILTER WHEEL, CO2  
AKIT, EXPENDABLES, M300E  
AKIT, TFE FLTR ELEM, 47MM, (FL6) (100)  
AKIT, TFE FLTR, 47MM, (FL6) (30)  
ASSY, SHUT-OFF VALVE, M300E (KB)  
INPUT MIRROR, REPLICATED  
OUTPUT MIRROR, REPLICATED  
WINDOW, SAMPLE FILTER, 47MM (KB)  
ASSY, SAMPLE FILTER, 47MM, ANG BKT, 5UM  
ASSY, SAMPLE THERMISTOR, M300 BRASS  
MIRROR, OBJECTIVE, 8 PASS,360E,320E,300E  
MIRROR, FIELD, 8 PASS  
ASSY, SCRUBBER, CO2, CH47, M360  
PCA, SYNC DEMOD w/DETECTOR, M360E  
ASSY, HEATER, OPTICAL BENCH  
KIT, EXPENDABLE, CO2 SCRUBBER  
ORING, TFE RETAINER, SAMPLE FILTER  
ASSY, FAN REAR PANEL, E SERIES  
PCA, FLOW/PRESSURE  
PCA, RELAY BOARD, M300E  
ASSY, CPU, CONFIGURATION, "E" SERIES  
ASSY, PUMP W/FLOW CONTROL  
ASSY, PUMP, INT, E SERIES  
PCA, KEYBOARD, E-SERIES, W/V-DETECT  
ASSY, VALVE, FOR SAMPLE/CAL VALVE ASSY  
ASSY, VALVE , SHUT-OFF  
ASSY, PUMP CONFIG PLUG, 100-115V/60 HZ  
PROGRAMMED FLASH, E SERIES  
ASSY, PWR CONF, 100-120V/60HZ, M3XXEX  
OPTION, PWR CONF, 220-240V/50HZ, M3XXEX  
OPTION, PWR CONF, 220-240V/60HZ, M3XXEX  
PCA, INTERFACE, ETHERNET, E-SERIES  
ASSY, CO2 SENSOR, M360E  
MANUAL, OPERATORS, M360E  
ASSY, PURGE REGULATOR, M360E  
PCA, SERIAL INTERFACE, w/ MD, E SERIES  
THERMAL PAD, DETECTOR HEATSINK  
05390H - M360E SPL (DCN5220)  
Page 1 of 2  
12/10/08  
Download from Www.Somanuals.com. All Manuals Search And Download.  
M360E Spare Parts List  
Part Number  
050320000  
Description  
PCA, OPTO-INTERRUPTER, M300E  
052560000  
052830200  
053220100  
053230000  
055010000  
055100200  
058021100  
CN0000458  
PCA, OPTO-INTERRUPTER, M360EX/GFC7000E  
ASSY, MOTOR HUB, MR7, "E", 115V  
DOC, w/SOFTWARE, M360E  
AKIT, EXPENDABLES, M360E IZS (w OPT 42C)  
ASSY, MTR WHL HEATER w/THERM, 200W  
OPTION, PUMP ASSY, 240V *  
PCA, E-SERIES MOTHERBOARD, GEN 5-I  
CONNECTOR, REAR PANEL, 12 PIN  
CN0000520  
DS0000025  
FL0000001  
FL0000003  
FM0000004  
HW0000020  
HW0000036  
HW0000090  
HW0000101  
HW0000379  
KIT000219  
OP0000009  
OR0000001  
OR0000002  
OR0000025  
OR0000034  
OR0000039  
OR0000041  
OR0000047  
OR0000077  
OR0000088  
OR0000094  
OR0000096  
OR0000097  
PS0000011  
PS0000024  
PS0000025  
PU0000022  
RL0000015  
SW0000051  
SW0000059  
VA0000014  
WR0000008  
CONNECTOR, REAR PANEL, 10 PIN  
DISPLAY, E SERIES (KB)  
FILTER, SS  
FILTER, DFU  
FLOWMETER (KB)  
SPRING  
TFE TAPE, 1/4" (48 FT/ROLL)  
SPRING, PURGE ORIFICE  
ISOLATOR  
SEAL, PURGED MOTOR ASSY  
KIT, 4-20MA CURRENT OUTPUT (E SERIES)  
WINDOW, IR SOURCE  
ORING, FLOW CONTROL  
ORING, WHEEL HOUSING SEAL  
ORING, CO2 SCRUBBER  
ORING, INPUT & OUTPUT MIRRORS  
ORING, IR SOURCE/BENCH  
ORING, OBJECT & FIELD MIRRORS  
ORING, WHEEL HOUSING SEAL  
ORING, 2-018V  
ORING, DETECTOR  
ORING, SAMPLE FILTER  
ORING, WHEEL HOUSING SEAL  
ORING, PURGED MOTOR ASSY  
PWR SUPPLY, SW, +5V, +/-15V, 40W (KB)  
POWER SUPPLY COVER  
PWR SUPPLY, SW, 12V, 40W (KB)  
REBUILD KIT, FOR PU20 & 04241 (KB)  
RELAY, DPDT, (KB)  
SWITCH, POWER CIRC BREAK VDE/CE, w/RG(KB  
PRESSURE SENSOR, 0-15 PSIA, ALL SEN  
REGULATOR  
POWER CORD, 10A, 6'  
05390H - M360E SPL (DCN5220)  
Page 2 of 2  
12/10/08  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Recommended Spare Parts Stocking Levels  
Model 360E  
Recommended Spare Parts Stocking Level: Standard  
Units  
6-10  
1
Part Number  
003290500  
Description  
1
1
2-5  
1
11-20  
1
21-30  
2
ASSY, THERMISTOR, M300 BENCH  
009550500  
037250100  
040010000  
040030100  
041710000  
ASSY, SOURCE, M300E S/N>65  
1
1
1
1
2
1
1
1
1
1
2
2
2
2
2
2
1
1
2
1
1
1
1
1
1
2
2
2
3
2
2
3
2
3
1
2
3
1
2
1
1
1
1
2
2
2
ASSY, BAND HEATER W/TC, M300EM/M3X0E  
ASSY, FAN REAR PANEL, E SERIES  
PCA, PRESS SENSORS (1X), w/FM4, E SERIES  
ASSY, CPU, CONFIGURATION, "E" SERIES  
ASSY, PUMP, INT, SOX/O3/IR  
042410200 *  
042580000  
052560000  
052840200  
055010000  
058021100  
DS0000025  
KIT000159  
KIT000180  
KIT000183  
PS0000011  
PS0000025  
RL0000015  
PCA, KEYBOARD, E-SERIES, W/V-DETECT  
PCA, OPTO-INTERRUPTER, M360EX/GFC7000E  
ASSY, MOTOR HUB, MR7, PURGED, "E", 115V  
ASSY, MTR WHL HEATER w/THERM, 200W  
PCA, E-SERIES MOTHERBOARD, GEN 5-I  
DISPLAY, E SERIES (KB)  
REPLACEMENT, RELAY BD, M300E, SN >= 100  
RETROFIT, SYNC DMOD UPDATE, M360E  
REPLACE, CO2 FILTER WHEEL ASSY, E-SERIES  
PWR SUPPLY, SW, +5V, +/-15V, 40W (KB)  
PWR SUPPLY, SW, 12V, 40W (KB)  
1
1
1
1
RELAY, DPDT, (KB)  
1
* Recommended Spare Parts Stocking Level: For Pump Assembly, 240V Option Installed  
Units  
6-10  
1
Part Number  
055100200  
Description  
1
2-5  
11-20  
21-30  
OPTION, PUMP ASSY, 240V  
2
2
Recommended Spare Parts Stocking Level: For IZS or ZS Option Installed  
Units  
6-10  
Part Number  
Description  
ASSY, VALVE (SS), M300E  
1
2-5  
11-20  
1
1
21-30  
2
2
042680000  
042690000  
ASSY, VALVE , 2-WAY, 12V  
04411M - M360E RSSL.xls (DCN 5289)  
02/25/09  
Download from Www.Somanuals.com. All Manuals Search And Download.  
THIS PAGE IS INTENTIONALLY LEFT BLANK  
Download from Www.Somanuals.com. All Manuals Search And Download.  
TELEDYNE  
Warranty/Repair  
Questionnaire  
Model 360E  
Instruction Manual  
INSTRUMENTS  
Advanced Pollution Instrumentation  
A Teledyne Technologies Company  
CUSTOMER: ____________________________________  
CONTACT NAME: ________________________________  
PHONE: ______________________________________  
FAX NO: ______________________________________  
SITE ADDRESS: __________________________________________________________________________________  
SERIAL NO.: ____________________________________ FIRMWARE REVISION: __________________________  
1. Are there any failure messages? ____________________________________________________________________  
________________________________________________________________________________________________  
________________________________________________________________________________________________  
________________________________________________________________________________________________  
________________________________________________________________________________________________  
Please complete the following table:  
PARAMETER  
DISPLAYED AS  
OBSERVED VALUE  
UNITS  
PPM  
PPM  
mV  
NOMINAL RANGE  
0 -10, 0 - 2000 Ppm  
0.15 Ppm With Zero Air  
3600 – 4800 Mv  
1400 – 2000 Mv  
2.5 ± 0.02 W/ Zero Air  
-2”Ambient Absolute  
800 ± 10%  
Range  
Stability  
STABIL  
CO2 MEAS  
CO2 Measure  
mV  
CO2 Reference  
Measure/Reference Ratio  
Pressure  
CO2 REF  
MR RATIO  
INHG  
CC  
ºC  
PRES  
Sample Flow  
SAMP FL  
48 ± 4  
Sample Temp  
SAMPLE TEMP  
BENCH TEMP  
WHEEL TEMP  
BOX TEMP  
ºC  
48 ± 2  
Bench Temp  
ºC  
68 ± 2  
Wheel Temp  
ºC  
Ambient + 7 ± 10  
250 Mv TO 4750 Mv  
1.0 ± .3  
Box Temp  
mV  
.
Photo Drive  
PHT DRIVE  
SLOPE  
Slope  
0 ± 0.3  
Offset  
OFFSET  
O2 Sensor Temperature3  
Slope of O2 Measurement3  
Offset of O2 Measurement3  
Dark Cal Reference signal  
Dark Cal Measurement Signal  
O2 CELL TEMP  
O2 SLOPE  
O2 OFFSET  
REF DARK OFFSET  
MEAS DARK OFFSET  
mV  
mV  
125 ± 50 Mv.  
125 ± 50 Mv  
PPM  
40 ± 2 Ppm  
Electric Test  
Values are in the Signal I/O  
4096mv±2mv And Must Be  
Stable  
REF_4096_MV  
REF_GND  
0± 0.5 And Must Be Stable  
2. Have you performed a leak check and flow check? ______________________________________________________  
3. What is the sample flow & sample pressure with the sample in-let on rear of machine capped?  
SAMPLE FLOW _________________________ CC  
SAMPLE PRESSURE ______________________ IN-HG-A  
TELEDYNE API CUSTOMER SERVICE  
EMAIL: api-customerservice@teledyne.com  
PHONE: (858) 657-9800 - TOLL FREE: (800) 324-5190 - FAX: (858) 657-9816  
C-1  
05235 Rev B  
Download from Www.Somanuals.com. All Manuals Search And Download.  
TELEDYNE  
Warranty/Repair  
Questionnaire  
Model 360E  
Instruction Manual  
INSTRUMENTS  
Advanced Pollution Instrumentation  
A Teledyne Technologies Company  
3. What are the failure symptoms? ____________________________________________________________________  
________________________________________________________________________________________________  
________________________________________________________________________________________________  
________________________________________________________________________________________________  
4. What test have you done trying to solve the problem? ___________________________________________________  
________________________________________________________________________________________________  
________________________________________________________________________________________________  
________________________________________________________________________________________________  
________________________________________________________________________________________________  
5. Please check these signals and verify the correctness. Look for the signals annotated on the diagram. What are the  
peak-to-peak voltages?  
TP 5  
TP 5  
TP 2  
2v/DIV  
10 mS  
2v/DIV  
.5 mS  
5. If possible, please include a portion of a strip chart pertaining to the problem. Circle pertinent data.  
THANK YOU FOR PROVIDING THIS INFORMATION. YOUR ASSISTANCE ENABLES TELEDYNE API TO RESPOND  
FASTER TO THE PROBLEM THAT YOU ARE ENCOUNTERING.  
OTHER INFORMATION: ____________________________________________________________________________  
________________________________________________________________________________________________  
________________________________________________________________________________________________  
________________________________________________________________________________________________  
________________________________________________________________________________________________  
________________________________________________________________________________________________  
________________________________________________________________________________________________  
________________________________________________________________________________________________  
________________________________________________________________________________________________  
________________________________________________________________________________________________  
________________________________________________________________________________________________  
________________________________________________________________________________________________  
________________________________________________________________________________________________  
TELEDYNE API CUSTOMER SERVICE  
EMAIL: api-customerservice@teledyne.com  
PHONE: (858) 657-9800 - TOLL FREE: (800) 324-5190 - FAX: (858) 657-9816  
C-2  
05235 Rev B  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model360E Instruction Manual  
APPENDIX D - ELECTRONIC SCHEMATICS  
APPENDIX D - ELECTRONIC SCHEMATICS  
Table D-1: List of Included Electronic Schematics  
DOCUMENT # DOCUMENT TITLE  
03297  
03632  
03976  
04003  
05703  
04089  
04136  
04216  
04217  
04259  
04468  
PCA, 03296, IR Photodetector Preamp and Sync Demodulator  
PCA, 03631, 0-20mA driver  
PCA, 03975, Keyboard & Display Driver  
PCA, 04003, Pressure/Flow Transducer Interface  
PCA, 05702, Motherboard, E-Series Gen 4  
PCA, 04088, Opto Pickup Interface  
PCA, 04135 Rev A, M300E Relay  
Interconnect Drawing - M300E SNs >=100  
Interconnect List - M300E SNs >=100  
PCA, 04258, Keyboard & Display Driver  
PCA, 04467, Analog Output Series Res  
User Notes:  
05236 Rev C  
D-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
APPENDIX D - ELECTRONIC SCHEMATICS  
Model 360E Instruction Manual  
D-2  
05236 Rev C  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Telex Musical Instrument FMR 500 User Manual
Toshiba Personal Computer SATELLITE PRO S200 User Manual
Toshiba Telephone CTX670 User Manual
Tote Vision Computer Monitor LCD 703HD1 User Manual
Tricity Bendix Oven SIE 250 User Manual
Turbosound Speaker TQ 308DP User Manual
Vanguard Heating Indoor Fireplace WMH26TNB User Manual
Vector Power Supply VEC020 User Manual
VocoPro DVD Player DVG 777K User Manual
Volvo MP3 Player Accessories 121 User Manual