Sanyo Freezer MDF U5486S User Guide

INSTRUCTION MANUAL  
MDF-U7486S  
MDF-U7486SC  
MDF-U5486S  
MDF-U5486SC  
Ultra-Low Temperature Freezer  
MDF-U7486S  
- 42 -  
Download from Www.Somanuals.com. All Manuals Search And Download.  
INTRODUCTION  
ꢀ Read this manual carefully before using the appliance and follow the instructions for safety operation.  
ꢀ Sanyo never guarantee any safety if the appliance is used for any objects other than intended use or  
used by any procedures other than those mentioned in this manual.  
ꢀ Keep this manual in an adequate place to refer to it as necessary.  
ꢀ The contents of the manual will be subjected to change without notice due to the improvement of  
performance or functions.  
ꢀ Contact Sanyo sales representative or agent if any page of the manual is lost or page order is incorrect.  
ꢀ Contact Sanyo sales representative or agent if any point in this manual is unclear or if there are any  
inaccuracies.  
ꢀ No part of this manual may be reproduced in any form without the expressed written permission of  
Sanyo.  
CAUTION  
SANYO guarantees the product under certain warranty conditions. SANYO in no way shall be  
responsible for any loss of content or damage of content.  
- 44 -  
2
Download from Www.Somanuals.com. All Manuals Search And Download.  
PRECAUTIONS FOR SAFE OPERATION  
It is imperative that the user complies with this manual as it contains  
important safety advice.  
Items and procedures are described so that you can use this unit correctly and safely.  
If the precautions advised are followed, this will prevent possible injury to the user and  
any other person.  
Precautions are illustrated in the following way:  
WARNING  
Failure to observe WARNING signs could result in a hazard to personnel  
possibly resulting in serious injury or death.  
CAUTION  
Failure to observe CAUTION signs could result in injury to personnel and  
damage to the unit and associated property.  
Symbol shows;  
this symbol means caution.  
this symbol means an action is prohibited.  
this symbol means an instruction must be followed.  
Be sure to keep this manual in a place accessible to users of this unit.  
Some warning and/or caution labels are attached on the unit. Following shows the description of such  
labels.  
This label is on the cover in which the electrical components of high voltage are  
enclosed to prevent the electric shock. The cover should be removed by a qualified  
engineer or a service personnel only.  
This symbol means attention or refer to document.  
This symbol means earth.  
This symbol means power switch “ON”.  
This symbol means power switch “OFF”.  
WARNING  
As with any equipment that uses CO2 gas, there is a likelihood of oxygen depletion in the vicinity  
of the equipment. It is important that you assess the work site to ensure there is suitable and  
sufficient ventilation. If restricted ventilation is suspected, then other methods of ensuring a safe  
environment must be considered. These may include atmosphere monitoring and warning  
devices.  
- 45 -  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
PRECAUTIONS FOR SAFE OPERATION  
WARNING  
Do not use the unit outdoors. Current leakage or electric shock may result if the unit is exposed to  
rain water.  
Only qualified engineers or service personnel should install the unit. The installation by  
unqualified personnel may cause electric shock or fire.  
Install the unit on a sturdy floor and take an adequate precaution to prevent the unit from  
turning over. If the floor is not strong enough or the installation site is not adequate, this may result  
in injury from the unit falling or tipping over.  
Never install the unit in a humid place or a place where it is likely to be splashed by water.  
Deterioration of the insulation may result which could cause current leakage or electric shock.  
Never install the unit in a flammable or volatile location. This may cause explosion or fire.  
Never install the unit where acid or corrosive gases are present as current leakage or electric  
shock may result due to corrosion.  
Always ground (earth) the unit to prevent electric shock. If the power supply outlet is not  
grounded, it will be necessary to install a ground by qualified engineers.  
Never ground the unit through a gas pipe, water main, telephone line or lightning rod. Such  
grounding may cause electric shock in the case of an incomplete circuit.  
Connect the unit to a power source as indicated on the rating label attached to the unit. Use  
of any other voltage or frequency other than that on the rating label may cause fire or electric shock.  
Never store volatile or flammable substances in this unit if the container cannot be sealed. These  
may cause explosion or fire.  
Do not insert metal objects such as a pin or a wire into any vent, gap or any outlet on the unit.  
This may cause electric shock or injury by accidental contact with moving parts.  
Use this unit in safe area when treating the poison, harmful or radiate articles. Improper use  
may cause bad effect on your health or environment.  
Turn off the power switch (if provided) and disconnect the power supply to the unit prior to any  
repair or maintenance of the unit in order to prevent electric shock or injury.  
Do not touch any electrical parts (such as power supply plug) or operate switches with a wet  
hand. This may cause electric shock.  
- 46 -  
4
Download from Www.Somanuals.com. All Manuals Search And Download.  
PRECAUTIONS FOR SAFE OPERATION  
WARNING  
Ensure you do not inhale or consume medication or aerosols from around the unit at the time of  
maintenance. These may be harmful to your health.  
Never splash water directly onto the unit as this may cause electric shock or short circuit.  
Never put containers with liquid on the unit as this may cause electric shock or short circuit when  
the liquid is spilled.  
Never bind, process, or step on the power supply cord, or never damage or break the power  
supply plug. A broken supply cord or plug may cause fire or electric shock.  
Do not use the supply cord if its plug is loose. Such supply cord may cause fire or electric shock.  
Never disassemble, repair, or modify the unit yourself. Any such work carried out by an  
unauthorized person may result in fire, or electric shock or injury due to a malfunction.  
Disconnect the power supply plug if there is something wrong with the unit. Continued  
abnormal operation may cause electric shock or fire.  
When removing the plug from the power supply outlet, grip the power supply plug, not the cord.  
Pulling the cord may result in electric shock or fire by short circuit.  
Disconnect the power supply plug before moving the unit. Take care not to damage the power  
cord. A damaged cord may cause electric shock or fire.  
Disconnect the power plug when the unit is not used for long periods. Keeping the connection  
may cause electric shock, current leakage, or fire due to the deterioration of insulation.  
If the unit is to be stored unused in an unsupervised area for an extended period, ensure that  
children do not have access and that doors cannot be closed completely.  
The disposal of the unit should be accomplished by appropriate personnel. Remove doors to  
prevent accidents such as suffocation.  
Do not put the packing plastic bag within reach of children as suffocation may result.  
- 47 -  
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
PRECAUTIONS FOR SAFE OPERATION  
CAUTION  
Use a dedicated power source (a dedicated circuit with a breaker) as indicated on the rating label  
attached to the unit. A branched circuit may cause fire resulting from abnormal heating.  
Connect the power supply plug to the power source firmly after removing the dust on the plug.  
A dusty plug or improper insertion may cause a heat or ignition.  
Never store corrosive substances such as acid or alkali in this unit if the container cannot be  
sealed. These may cause corrosion of inner components or electric parts.  
Check the setting when starting up of operation after power failure or turning off of power  
switch. The stored items may be damaged due to the change of setting.  
Be careful not to tip over the unit during movement to prevent damage or injury.  
Prepare a safety check sheet when you request any repair or maintenance for the safety of service  
personnel.  
- 48 -  
6
Download from Www.Somanuals.com. All Manuals Search And Download.  
ENVIRONMENTAL CONDITIONS  
This equipment is designed to be safe at least under the following conditions (based on the IEC 61010-1):  
ꢀ Indoor use;  
ꢀ Altitude up to 2000 m;  
ꢀ Ambient temperature 5oC to 40oC  
ꢀ Maximum relative humidity 80% for temperature up to 31oC decreasing linearly to 50% relative humidity  
at 40oC;  
ꢀ Mains supply voltage fluctuations not to exceed 10% of the nominal voltage;  
ꢀ Other supply voltage fluctuations as stated by the manufacturer;  
ꢀ Transient overvoltages according to Installation Categories (Overvoltage Categories) II; For mains  
supply the minimum and normal category is II;  
ꢀ Pollution degree 2 in accordance with IEC 60664.  
- 49 -  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
FREEZER COMPONENTS  
13  
1
2
3
14  
4
13 (inside)  
12  
5
6
7
15  
9 (inside)  
8
11  
16  
10  
17  
18  
19  
Power cord  
Rear side  
- 50 -  
8
Download from Www.Somanuals.com. All Manuals Search And Download.  
FREEZER COMPONENTS  
1. Outer door: To open the outer door, grip the handle. On closing, lock the outer door latch  
completely.  
2. Inner door: The operation of the inner door should be quick to minimize the temperature rise in  
chamber. Lock the inner door latch completely when the inner door is closed. The inner door is  
removable for cleaning or defrosting. See page 21 “Routine maintenance”.  
3. Inner door latch: Always lock the inner door latch when the inner door is closed.  
4. Outer door latch: Always lock the latch when the outer door is closed.  
5. Magnetic door gasket: This provides a tight door seal and prevents cold air leak. Keep clean.  
6. Air intake vent (grille): Do not block this vent to keep the proper cooling performance.  
7. Caster: 4 casters are provided to facilitate moving of the cabinet. For the installation, adjust the  
leveling foot so that the front 2 casters cannot contact with the floor.  
8. Leveling foot: These are screw bolts used to install and fix the unit. Adjust the height of the  
leveling feet by turning the screw bolts until 2 front casters are away from the floor.  
9. Condenser filter (behind the grille): This filter prevents the dust from accumulating on the  
condenser. The dusty filter may cause failure of refrigerating device. Clean the filter once a month.  
See page 20 “Routine maintenance” for the cleaning.  
10. Space for temperature recorder: A temperature recorder (optional component) can be attached  
here. See page 27 “Temperature recorder”.  
11. Lock: Turn clockwise to 180o with a key and the outer door is securely locked.  
12. Control panel: Used for temperature setting and indication of operating status is displayed on the  
panel. See page 10 for details.  
13. Access port (rear and bottom): This is used for leading a cable and sensor of a measuring  
equipment, or nozzle of back-up system to chamber.  
14. Air intake port: After closing the outer door, if used to open soon. See page 17  
15. Fixture (on back side): 2 fixtures serve as spacers between the cabinet and wall and also serve as  
hooks to fix the unit. See page 12 “Installation”.  
16. Power switch: This is for turning ON/OFF the power to the unit. ON – “I” OFF – “”  
17. Remote alarm terminal: This is used to notice an alarm condition of the unit to remote location.  
Refer to page 17 “Remote alarm terminal”.  
18. Battery switch: This is a switch for a battery for power failure alarm. Normally, turn on this switch.  
Be sure to turn off this switch if the freezer is not in operating for the long period.  
19. Fuse: 125 V, 3.15 A. Normal blow type.  
- 51 -  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
FREEZER COMPONENTS  
Control panel  
2
1
3
4
7
6
9
8
5
1. Digital temperature indicator: This indicator shows the present chamber temperature or set  
temperature.  
2. Battery check lamp (BATTERY): This lamp flickers to recommend the battery replacement. For  
the replacement, consult Sanyo sales representative or agent.  
3. Alarm lamp (ALARM): This lamp is flashed during alarm condition.  
4. Filter lamp (FILTER): This lamp lights when the excessive dust is accumulated on the condenser  
filter. When this lamp lights, clean the condenser filter following the procedure on page 20.  
5. Numerical value shift key (  
): Pressing this key in the setting mode causes the numerical value to  
shift. “ON-OFF” of key lock can be selected by pressing this key in the key lock mode. By pressing this  
key for more than 5 seconds in the temperature display mode leads setting mode for alarm temperature,  
alarm resume time ,compressor delay time and door alarm delay time. Refer to page 15 and 16 for  
details respectively.  
6. Digit shift key (  
): Pressing this key in the setting mode causes the changeable digit to shift. Key  
lock is available by pressing this key for more than 5 seconds in the temperature display mode. Refer to  
page 14 for details.  
7. Set key (SET): Temperature set mode is led by pressing this key and the changeable digit is flashed.  
By pressing this key again, the setting is memorized. The set mode returns to the temperature display  
mode automatically when 90 seconds has passed without any key operation. Refer to page 14 for  
details.  
8. Alarm test key (ALARM TEST): To check the alarm system. Pressing this key with the battery  
switch ON gets the alarm lamp to flash, the remote alarm to operate, and the buzzer to sound.  
9. Alarm buzzer stop key (BUZZER): To silence the audible alarm under alarm condition, press this  
key. The buzzer during alarm test cannot be silenced by this key.  
- 52 -  
10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
INSTALLATION SITE  
To operate this unit properly and to obtain maximum performance, install the unit in a location with the  
following conditions:  
A location not subjected to direct sunlight  
Do not install the unit under direct sunlight. Installation in a location subjected to direct sunlight cannot  
obtain the intended performance.  
A location with adequate ventilation  
Leave at least 10 cm around the unit for ventilation. Poor ventilation will result in a reduction of the  
performance and consequently the failure.  
A location away from heat generating sources  
Avoid installing the unit near heat-emitting appliances such as a heater or a boiler etc. Heat can  
decrease the intended performance of the unit.  
A location with little temperature change  
Install the unit under stable ambient temperature. The allowable ambient temperature is between +5  
and +30oC.  
A location with a sturdy and level floor  
Always install the unit on a sturdy and level floor. The uneven floor or tilted installation may cause failure  
or injury. Install the unit in stable condition to avoid the vibration or noise. Unstable condition may  
cause vibration or noise.  
WARNING  
Install the unit on a sturdy floor. If the floor is not strong enough or the installation site is not  
adequate, this may result in injury from the unit falling or tipping over.  
Select a level and sturdy floor for installation. This precaution will prevent the unit from tipping.  
Improper installation may result in water spillage or injury from the unit tipping over.  
A location not prone to high humidity  
Install the unit in the ambient of 80% R.H. or less humidity. Installation under high humidity may cause  
current leakage or electric shock.  
WARNING  
Do not use the unit outdoors. Current leakage or electric shock may result if the unit is exposed to  
rain water.  
Never install the unit in a humid place or a place where it is likely to be splashed by water.  
Deterioration of the insulation may result which could cause current leakage or electric shock.  
A location without flammable or corrosive gas  
Never install the unit in a flammable or volatile location. This may cause explosion or fire or may result  
in the current leakage or electric shock by the corrosion of the electrical components.  
A location without the possibility of anything fall  
Avoid installing the unit in the location where anything can fall down onto the unit. This may cause the  
breakdown or failure of the unit.  
- 53 -  
11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
INSTALLATION  
1. Remove the packaging materials and tapes  
Remove all transportation packaging materials and tapes. Open the doors and ventilate the unit. If the  
outside panels are dirty, clean them with a diluted neutral dishwashing detergent. (Undiluted detergent  
can damage the plastic components. For the dilution, refer to the instruction of the detergent.) After  
the cleaning with the diluted detergent, always wipe it off with a wet cloth. Then wipe off the panels with  
a dry cloth.  
2. Adjust the leveling foot  
Extend the leveling feet by rotating them  
counterclockwise to contact them to the floor.  
Ensure the unit is level. (Fig.1)  
Leveling foot  
Fig.1  
3. Fix the unit  
Two fixtures are attached to the rear of the frame.  
Fix the frame to the wall with these fixtures and rope  
or chain. (Fig.2)  
Fixture  
Fig.2  
WARNING  
Use a power supply outlet with ground (earth) to prevent electric shock. If the power supply outlet  
is not grounded, it is necessary to install a ground by qualified engineers.  
Never ground the unit through a gas pipe, water main, telephone line or lightning rod. Such  
grounding may cause electric shock in the case of an incomplete circuit.  
- 54 -  
12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
START-UP OF UNIT  
Follow the procedures for the initial and consequent operations of the unit.  
1. Turn off the switch of the back-up system (optional component) if it is installed.  
2. Connect the power supply cord to the dedicated outlet having appropriate rating with the chamber  
empty, and turn on the power switch on the freezer.  
3. Turn on the battery switch.  
4. Set the desired chamber temperature. See page 14 for the temperature setting.  
5. Check that the chamber temperature reaches the desired temperature.  
6. Turn on the switch of back-up system (optional component) if it is installed.  
7. Make sure that the alarm lamp blinks and the buzzer sounds by pressing the alarm test key (ALARM  
TEST). The remote alarm is also operated. E09 is displayed on the digital temperature indicator if the  
battery switch is OFF. Make sure to turn on the battery switch.  
8. After confirming the above, you can put articles into the freezer chamber in a small batch to prevent the  
temperature rise.  
9. An alarm function is to work normally, or push and confirm an alarm test key (the condition that a  
battery switch is on). It is normal if alarm lamp blinking, remote alarm operation and a buzzer ring.  
Operation after power failure  
The set value is memorized by nonvolatile memory. Accordingly, the freezer resumes the operation with  
setting before power failure.  
When the freezer is recovered from power failure with the chamber temperature higher than the preset  
temperature, then the high temperature alarm is activated and the buzzer sounds and the remote alarm is  
also activated. Please push the buzzer stop key (BUZZER) to silence buzzer and take appropriate  
actions if needed.  
- 55 -  
13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
CHAMBER TEMPERATURE SETTING  
Table 1 shows the basic procedure for setting the chamber temperature. Perform key operations in the  
sequence indicated in the table. The example in the table is based on the assumption that the desired  
temperature is -75oC.  
Note: The unit is set at the factory that the chamber temperature -80oC.  
Table 1. Basic operation sequence (Example: Chamber temperature -75oC)  
Description of operation  
Key operated  
Indication after operation  
The current chamber temperature is  
displayed.  
1
2
Turn the power switch ON.  
----  
Press set key.  
SET  
The second digit is flashed.  
When pressed, the figure of settable  
digit changes.  
When pressed, the settable digit  
is shifted.  
Set temperature is memorized and the  
current chamber temperature is  
displayed.  
Set to -75 with the numerical value  
shift key and digit shift key.  
3
4
Press set key.  
SET  
Note:  
• The temperature set mode returns to the temperature display mode automatically when 90 seconds has  
passed without any key operation.  
• Although the value of the chamber temperature setting can range from -50oC to -90oC, the guaranteed  
temperature when there is no load is -85oC when the ambient temperature is 30oC.  
• The set mode returns to the temperature display mode automatically when 90 seconds has passed  
without any key operation. In this case, any setting before pressing set key (SET) is not memorized.  
KEY LOCK FUNCTION  
This unit is provided with the key lock function. When the key lock is ON, change of temperature setting  
through the key pad is not available. The key lock is set in OFF at the factory.  
Display  
Mode  
Key lock is OFF  
Key lock is ON  
Function  
L
L
0
1
Enable to change of temperature setting  
Disable to change of temperature setting  
Table 2. Procedure for key lock setting (change from key lock OFF to key lock ON)  
Description of operation  
Key operated  
Indication after operation  
The current chamber temperature is  
displayed.  
1
2
3
----  
Press digit shift key for 5 seconds.  
The first digit is flashed.  
Press numerical value shift key  
and scroll the figure to 1.  
When pressed, the figure of settable  
digit changes.  
The key lock is set to ON.  
The current chamber temperature is  
displayed.  
4
Press set key.  
SET  
• The set mode returns to the temperature display mode automatically when 90 seconds has passed  
without any key operation. In this case, any setting before pressing set key (SET) is not memorized.  
- 56 -  
14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
ALARM TEMPERATURE SETTING  
This unit is provided with the high and low temperature alarm and the temperature at which the alarm is  
activated is changeable.  
The following procedure shows the setting of alarm temperature according to the condition below:  
High temperature alarm: activates at the temperature 5oC higher than the set temperature  
Low temperature alarm: activates at the temperature 5oC lower than the set temperature  
Note:  
The alarm temperature is set at the factory 10oC higher and lower than the set temperature.  
The available range of alarm temperature is between 5oC and 40oC higher or lower than the set  
temperature.  
Table 3. Procedure for setting high temperature alarm  
Description of operation  
Key operated  
Indication after operation  
The current chamber temperature is  
displayed.  
1
2
3
4
----  
Press numerical value shift key for  
about 5 seconds.  
Press numerical value shift key  
and scroll the figure to 1.  
The first digit is flashed.  
The first digit is flashed.  
The first digit is flashed.  
Press set key.  
SET  
SET  
When pressed, the figure of settable  
digit changes.  
When pressed, the changeable digit  
moves.  
Alarm temperature is memorized and  
the current chamber temperature is  
displayed.  
Scroll the figure to 005 by using  
digit shift key and numerical value  
shift key.  
5
6
Press set key.  
Table 4. Procedure for setting low temperature alarm  
Description of operation  
Key operated  
Indication after operation  
The current chamber temperature is  
displayed.  
1
2
3
4
----  
Press numerical value shift key for  
about 5 seconds.  
Press numerical value shift key  
and scroll the figure to 2.  
The first digit is flashed.  
The first digit is flashed.  
The first digit is flashed.  
Press set key.  
SET  
When pressed, the figure of settable  
digit changes.  
When pressed, the changeable digit  
moves.  
Scroll the figure to -05 by using  
digit shift key and numerical value  
shift key.  
5
Alarm temperature is memorized and  
the current chamber temperature is  
displayed.  
6
Press set key.  
SET  
• The set mode returns to the temperature display mode automatically when 90 seconds has passed  
without any key operation. In this case, any setting before pressing set key (SET) is not memorized.  
- 57 -  
15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SETTING OF ALARM RESUME TIME  
The alarm buzzer is silenced by pressing alarm buzzer stop key (BUZZER) on the control panel during  
alarm condition. (A remote alarm is continuing the operation.)  
The buzzer will be activated again after certain suspension if the alarm condition is continued. The  
suspension time can be set by following the procedure shown in the Table 5 below.  
The example in the table is based on the assumption that the desired duration is 20 minutes.  
Note: The duration is set in 30 minutes at the factory.  
Table 5. Setting procedure for alarm resuming time (change from 30 minutes to 20 minutes)  
Description of operation  
Key operated  
Indication after operation  
The current chamber temperature is  
displayed.  
1
2
----  
Press digit shift key for 5 seconds.  
The first digit is flashed.  
The settable digit is shifted.  
Set the figure to F25 with the digit  
shift key and numerical value shift  
key.  
3
When pressed, the figure of settable  
digit changes.  
The current resume time is displayed.  
The second digit is flashed.  
4
5
Press set key.  
SET  
SET  
Set the figure to 020 with the  
numerical value shift key.  
When pressed, the figure of settable  
digit changes.  
The setting is memorized and  
the current chamber temperature is  
displayed.  
6
Press set key.  
• The settable alarm resume time is 10, 20, 30, 40, 50, or 60 minutes (The setting is 010, 020, 030, 040,  
050, or 060). The buzzer would not reset if the reset time is set in 000.  
• It is recommended to set the alarm resume time when the freezer is not under alarm condition. The  
setting during alarm condition is effective on the next alarm condition.  
• The setting cannot be changed during power failure.  
• The set mode returns to the temperature display mode automatically when 90 seconds has passed  
without any key operation. In this case, any setting before pressing set key (SET) is not memorized.  
- 58 -  
16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
REMOTE ALARM TERMINAL  
The terminal of the remote alarm is installed at the lower left side of the unit. The alarm is outputted from  
this terminal. Contact capacity is DC 30V, 2 A.  
Contact output:  
between COM. and N.O.  
between COM. and N.C.  
At normal  
At abnormal  
Open  
Close  
Close  
Open  
Note:  
• The alarm buzzer is silenced by pressing alarm buzzer stop key (BUZZER) on the control panel during  
alarm condition. (A remote alarm is continuing the operation.)  
The buzzer will be activated again after certain suspension if the alarm condition is continued.  
• The alarm is actuated when the power supply plug is disconnected from the outlet or the power switch is  
OFF.  
REMOTE ALARM  
N.C. COM N.O.  
MAX DC30V 2A  
AIR INTAKE PORT  
When a door is closed and opened soon, a door does not  
sometimes open this product.  
The warm air which went into the chamber is cooled down rapidly,  
and this is because air inside the chamber contracted.  
When a door is closed and opened soon, a door is easy to open  
with the following process.  
1. Turn the cap of the left side counterclockwise, and remove it.  
2. Put the air into the chamber from about ten for twenty seconds,  
and open a door.  
Cap  
3. Close a cap if a door opens.  
A door may not open in the above method when there are frost and ice in the air intake port. In that  
case, open a cap, and investigate the matter whether there is no frost inside the air intake port. Remove  
frost inside the air intake port in "the air intake port cleaning stick" of the accessories when there is frost.  
Clean a air intake port every month even when there is no frost inside the air intake port. Refer to page  
22.  
WARNING  
For removing the frost of the air intake port, do not use a tool with sharp edge such as a knife or a screw  
driver.  
- 59 -  
17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SETTING OF COMPRESSOR DELAY TIME  
The delay time of high and low stage side compressor can be changed to reduce the load on the power  
line and to facilitate the start-up (reset) of the freezer after power failure.  
The example in the table is based on the assumption that the delay time is changed to 4 minutes.  
(The delay time is set in 3 minutes at the factory.)  
Note:  
• The setting range for delay time is between 3 and 15 minutes. The cool down of chamber temperature  
may be slow when the setting of delay time is over 5 minutes, depending on the installation environment.  
There is no need of changing the delay time when the capacity of power source is adequate.  
Table 6. Setting procedure for delay time (change from 3 minutes to 4 minutes)  
Description of operation  
Key operated  
Indication after operation  
The current chamber temperature is  
displayed.  
1
2
----  
Press numerical value shift key for  
5 seconds.  
The first digit is flashed.  
Set the figure to F05 with the  
numerical value shift key.  
When pressed, the figure of settable  
digit changes.  
The current delay time is displayed.  
The first digit is flashed.  
3
4
5
Press set key.  
SET  
SET  
Set the figure to 004 with the  
numerical value shift key.  
When pressed, the figure of the first  
digit changes.  
The delay time is memorized and  
the current chamber temperature is  
displayed.  
6
Press set key.  
• The compressor starts to operate with the delay time set by the above procedure at the time of power on  
or after power failure. However, the start up of the low stage side compressor is affected by the  
chamber temperature and the cascade condenser temperature. The delay time varies depending on  
how they meet the start up conditions.  
• The set mode returns to the temperature display mode automatically when 90 seconds has passed  
without any key operation. In this case, any setting before pressing set key (SET) is not memorized.  
- 60 -  
18  
Download from Www.Somanuals.com. All Manuals Search And Download.  
ALARMS & SAFETY FUNCTIONS  
This unit has the alarms and safety functions shown in Table 7, and also self diagnostic functions.  
Table 7. Alarms and safety functions  
Alarm  
Situation  
Indication  
Buzzer  
Safety functions  
If the chamber temperature is higher  
than the temperature at which the  
high temperature alarm is activated.  
If the chamber temperature is lower  
than the temperature at which the  
low temperature alarm is activated.  
When the power to the unit is  
disconnected.  
When the condenser filter is clogged.  
When about 3 years has passed with  
power switch ON.  
High temperature  
alarm  
Alarm lamp is flashed.  
Temperature indicator is  
flashed.  
Intermittent tone with  
15 minutes delay.  
Remote alarm with 15  
minutes delay.  
Low temperature  
alarm  
Power failure alarm  
Filter alarm  
Alarm lamp is flashed.  
Filter lamp lights.  
Intermittent tone  
Intermittent tone  
-----  
Remote alarm.  
-----  
-----  
Battery check  
Battery check lamp lights.  
When about 6 years has passed with  
Battery check lamp  
flashes.  
Chamber temperature is  
displayed.  
Fan motor check  
Auto-return  
Key lock  
-----  
-----  
-----  
power switch ON.  
When there is no key pressing in  
each set mode for 90 seconds.  
Finishing of each  
set mode.  
Change of setting is  
When the key lock is ON.  
-----  
-----  
disable.  
Alarm lamp is flashed.  
E01 and chamber temp. are  
displayed alternately.  
Remote alarm.  
Unit keeps continuous  
running.  
If the thermal sensor is disconnected.  
Intermittent tone  
Alarm lamp is flashed.  
If the thermal sensor is short-circuited. E02 and chamber temp. are  
displayed alternately.  
Alarm lamp is flashed.  
If the cascade sensor is disconnected. E03 and chamber temp. are  
displayed alternately.  
Remote alarm.  
Unit keeps continuous  
running.  
Intermittent tone  
Intermittent tone  
Intermittent tone  
Intermittent tone  
Intermittent tone  
Intermittent tone  
Remote alarm.  
Remote alarm.  
Remote alarm.  
Remote alarm.  
Remote alarm.  
Alarm lamp is flashed.  
If the cascade sensor is short  
E04 and chamber temp. are  
circuited.  
displayed alternately.  
Alarm lamp is flashed.  
Sensor abnormality  
If the filter sensor is disconnected.  
If the filter sensor is short-circuited.  
E05 and chamber temp. are  
displayed alternately.  
Alarm lamp is flashed.  
E06 and chamber temp. are  
displayed alternately.  
Alarm lamp is flashed.  
E07 and chamber temp. are  
displayed alternately.  
If the ambient temperature sensor is  
disconnected.  
Alarm lamp is flashed.  
E08 and chamber temp. are  
displayed alternately.  
If the ambient temperature sensor is  
short-circuited.  
Intermittent tone  
-----  
Remote alarm.  
-----  
Battery switch  
check  
When the battery switch is OFF during Alarm lamp is flashed.  
alarm test.  
E09 is flashed.  
Remote alarm.  
Compressor of high  
stage side stops.  
Condenser temp.  
abnormality  
In the event of failure of fan motor for  
cooling the compressor  
E10 and chamber temp.  
are displayed alternately.  
Intermittent tone  
Note:  
• When the operation is started in high ambient temperature, the filter lamp is sometimes flashed. In this  
case, the lamp is off automatically when the chamber temperature is getting lower.  
• The freezer resumes the operation after power failure with the temperature setting before power failure  
as the chamber temperature setting and alarm temperature setting are memorized in the nonvolatile  
memory.  
• The battery for power failure alarm is an article for consumption. It is recommended that the battery  
will be replaced about every 3 years. Contact Sanyo sales representative or agent at the time of  
replacement of the battery for recycling.  
• Fan motors are expendable supplies. Replace them for about every 6 years. Contact Sanyo sales  
representative or agent at the time of replacement of the fan motor.  
• The chamber temperature is displayed for 5 seconds by pressing alarm buzzer stop key (BUZZER)  
during power failure alarm. The alarm lamp keeps flashing.  
- 61 -  
19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
ROUTINE MAINTENANCE  
WARNING  
Always disconnect the power supply to the unit prior to any repair or maintenance of the unit in  
order to prevent electric shock or injury.  
Ensure you do not inhale or consume medication or aerosols from around the unit at the time of  
maintenance. These may be harmful to your health.  
Cleaning of cabinet  
Clean the unit once a month. Regular cleaning keeps the unit looking new.  
Use a dry cloth to wipe off small amounts of dirt on the outside and inside of the unit and all  
accessories. If the outside panels are dirty, clean them with a diluted neutral dishwashing detergent..  
(Undiluted detergent can damage the plastic components. For the dilution, refer to the instruction of the  
detergent.)  
After the cleaning with the diluted detergent, always wipe it off with a wet cloth. Then wipe off the  
cabinet or accessories with a dry cloth.  
Never pour water onto or into the unit. Doing so can damage the electric insulation and cause failure.  
The compressor and other mechanical part are completely sealed. This unit requires absolutely no  
lubrication.  
Remove the frost or ice on the chamber wall and clean the condenser filter once a month.  
Cleaning of condenser filter  
This unit is provided with the filter lamp on the control panel. Clean the condenser filter when this lamp  
lights. Clean the filter once a month even if the lamp is not on since a clogged condenser filter may  
cause shorter compressor life as well as the poor cooling.  
Clean the condenser filter by the procedure below.  
1. Open the grille by pulling it to you as shown in the  
figure.  
Condenser  
filter  
2. Take out the condenser filter.  
Grille  
3. Wash the condenser filter with water.  
4. Replace the condenser filter and the grille.  
5. Check that the filter lamp is off in the event the lamp  
was ON.  
WARNING  
Do not touch the condenser directly when the condenser filter is removed for cleaning. This may  
cause injury by hot surface.  
- 62 -  
20  
Download from Www.Somanuals.com. All Manuals Search And Download.  
ROUTINE MAINTENANCE  
Defrosting of inside wall  
The frost is built at the upper portion of the chamber and inner door. The excessive frost possibly make  
some gap between the cabinet and magnetic door gasket, which may cause poor cooling. Remove the  
frost on the inner door with a scraper enclosed with the unit. Following shows the procedure for  
removing the chamber frost.  
Note: For removing the frost, do not use a tool with sharp edge such as a knife or a screw driver.  
1. Turn off the back-up system if applicable.  
2. Take out and transfer all the contents to another  
freezer or a container which is refrigerated by liquid  
carbon dioxide or dry ice.  
3. Turn off the power switch of the freezer.  
4. Open the outer door and inner door. Remove the  
inner door by lifting up as shown in the figure.  
5. Leave the freezer as it is.  
6. The water accumulated on the bottom of the chamber  
should be wiped up with a dry cloth.  
7. After cleaning the chamber and inner door, replace  
the inner door and start up the unit according to the  
procedure on page 13.  
8. Put back the articles into the sufficiently cooled  
freezer compartment.  
9. Turn on the back-up system if it is provided.  
- 63 -  
21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
ROUTINE MAINTENANCE  
Cleaning of air intake port  
The cap of the left side is turned counterclockwise (Counterclockwise direction), and this product is  
removed, and the outer air is adopted into the chamber, and it opens a outer door. Therefore, frost is  
easy to be settled around the air intake port inside the chamber. Clean it in the case shown below.  
Condition  
Check / Remedy  
When frost and ice can be  
seen in the port of the air  
intake port.  
The pipe of the air intake port is thrust with a stick for air intake port  
cleaning of the accessories, and frost is taken.  
An outside door does not  
open even if the cap of the  
air intake port is opened.  
The pipe of the air intake port is thrust with a stick for air intake port  
cleaning of the accessories, and frost is taken.  
Frost and ice can be seen in Frost inside the chamber and ice are taken with Hera of the  
the chamber.  
accessories.  
Stick for air intake port cleaning (Accessory)  
WARNING  
For removing the frost of the release port, do not use a tool with sharp edge such as a knife or a screw  
driver.  
- 64 -  
22  
Download from Www.Somanuals.com. All Manuals Search And Download.  
TROUBLE SHOOTING  
If the unit malfunctions, check out the following before calling for service.  
Malfunction  
The chamber is not cooled  
at all  
Check/Remedy  
The circuit breaker of power source is active.  
The voltage is too low (In this case, call an electrician).  
The power is not supplied.  
The breaker is free.  
The large amount of articles (load) is stored in the chamber at one  
time.  
The ambient temperature is too high.  
The latch of inner door is not closed completely. The outer door  
is not closed firmly. (The frost or ice between the cabinet and door  
gasket possibly prevents door seal.)  
The cooling is poor  
The air intake vent is blocked.  
The condenser filter is clogged. Always clean the filter when the  
filter check lamp is lit.  
The door is not shut tightly.  
The inner door is not installed correctly.  
The set temperature in the controller is not set properly.  
The freezer is in the direct sunlight.  
There is any heating source near the freezer.  
A rubber cap and insulation are not set correctly.  
You put too many unfrozen articles into the chamber.  
The alarm is activated only when the power switch is ON.  
When only the buzzer or only the alarm is actuated by the alarm test  
key, the unactuated part is out of order, and must be replaced.  
Alarm test key cannot  
actuate the alarm  
Note:  
If the malfunction is not eliminated after checking the above items, or the malfunction is not shown in the  
above table, contact Sanyo sales representative or agent.  
If the unit malfunctions, check out the following before calling for service.  
- 65 -  
23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
DISPOSAL OF UNIT  
WARNING  
If the unit is to be stored unused in an unsupervised area for an extended period ensure that children do  
not have access and doors cannot be closed completely.  
The disposal of the unit should be accomplished by appropriate personnel. Always remove  
doors to prevent accidents such as suffocation.  
Recycle of battery  
The unit contains a rechargeable battery. The battery is recyclable. At the end  
of it’s useful life, check with you local solid officials option or proper disposal.  
ɼႱ΍ូ  
ਣƧ˂ូ  
Label indication is obliged to comply with Taiwanese battery regulation.  
Decontamination of unit  
Before disposal of unit, decontaminate the unit yourself.  
- 66 -  
24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
DISPOSAL OF BATTERY  
Location of a nickel-metal-hydride battery  
This unit is provided a nickel-metal-hydride battery for the power failure warning device. The battery is  
located in the electrical box inside the cover on the lower left side. (Fig. 1)  
The high voltage components are enclosed in the electrical box. The cover should be removed  
by a qualified engineer or a service personnel only to prevent the electric shock.  
Disposal of nickel-metal-hydride battery  
1. Turn off the power switch and disconnect the power supply plug.  
2. As shown in the Fig. 2, remove 6 screws fixing the side cover with a screw driver and remove the side  
cover.  
3. Remove 5 screws fixing the electrical box cover with a screw driver. (Fig. 3)  
4. Disconnect the battery connector and remove 2 screws fixing the battery mounting plate. (Fig. 4)  
5. Take out the battery.  
6. Follow the procedure for recycling or proper disposal.  
Switch cover  
Side cover  
Electrical box  
Fig. 2  
Fig. 1  
Electrical box cover  
Fig. 3  
Battery  
Fig. 4  
mounting plate  
- 67 -  
25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
BACK-UP SYSTEM (OPTION)  
WARNING  
As with any equipment that uses CO2 gas, there is a likelihood of oxygen depletion in the vicinity  
of the equipment. It is important that you assess the work site to endure there is suitable and  
sufficient ventilation. If restricted ventilation is suspected, then other methods of ensuring a safe  
environment must be considered. These may include atmosphere monitoring and warning  
devices.  
This freezer can be provided with a back-up system (CVK-UB2) which is available as an optional  
component. For the installation, refer to the instruction manual enclosed with the system.  
1. Switch of back-up system (BACKUP)  
When turning on the system, the lamp is brightened.  
This means that the system is ready. To stop the  
1
operation of the system, turn off this switch.  
2
2. Test switch (TEST)  
This switch is for checking the operation of back-up  
system. Pressing this switch is resulted in the release  
of liquid carbon dioxide without system operation.  
3
3. Temperature setting knob (TEMP. SET)  
With this knob, set the temperature at which the system  
is operated. The effective set temperature range is  
between -50oC and -70oC.  
- 68 -  
26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
TEMPERATURE RECORDER (OPTION)  
Installation of MTR-85H/MTR-G85  
Contact Sanyo sales representative or agent for the attachment of an automatic temperature recorder. :  
INTERFACE BOARD (OPTION)  
Installation of MTR-480/MTR-L03  
Contact Sanyo sales representative or agent for the attachment of a interface board. :  
Note:  
When a data transmitting function to the personal computer is done, an interface board MTR-480  
(option goods) and communication cable of 9 pin Dsub cross type for RS232C are necessary.  
The connecting cable between the PC and MTR-L03 terminal is not enclosed. Provide a cross type  
cable of less than 100 m length having a pin plug.  
- 69 -  
27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
INVENTORY RACK (OPTION)  
The optional inventory rack is useful to store the precious materials in the chamber effectively. When  
the rack is used, it is necessary to adjust the height of the shelves. Set the shelf support as shown in the  
figure below.  
IR-220U  
Shelf support  
MDF-U5486SC  
NO. 8 under the second  
screw from the top  
Shelf stopper  
No. 3 under the ffifth  
screw from the top  
No. 2 under the third  
screw from the bottom  
- 70 -  
28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SPECIFICATIONS  
Name  
Model  
Ultra-Low Temperature Freezer  
MDF-U5486S MDF-U5486SC  
External dimensions  
Internal dimensions  
Effective capacity  
Exterior  
W890 x D870 x H1990 (mm)  
W630 x D600 x H1280 (mm)  
483 L  
Painted steel  
Interior  
Painted steel  
Outer door  
Inner door  
Painted steel  
ABS resin panel with stainless frame, 2 doors  
Shelf  
Stainless steel, 3 shelves (adjustable)  
W608 x D535 (mm), Load; 50 kg/shelf  
Access port  
17 mm diameter, 2 locations (back, bottom left corner)  
Insulation  
Compressor  
Evaporator  
Rigid polyurethane foamed-in place  
Hermetic type, Output; 1100 W (high stage side), 1100 W (low stage side)  
Tube on sheet type  
Condenser  
Refrigerant  
Temperature controller  
Temperature display  
Thermal sensor  
Alarm  
Fin and tube type (high stage side), Shell and tube type (low stage side)  
R-404A (high stage side), R-508B (low stage side)  
Microcomputer control system  
Digital display  
Platinum resistance (Pt 1000)  
High temp. alarm, Low temp. alarm, Power failure alarm  
Filter check, Battery check, Fan motor check  
Allowable contact capacity: 30 V DC, 2 A  
Remote alarm contact  
Battery  
Accessories  
Weight  
Nickel-metal-hydride battery, 6 V DC, 1100 mAh, Auto-recharge  
1 set of key, 1 scraper, 1 stick for air intake port cleaning  
315 kg  
None  
320 kg  
Built-in  
Voltage booster  
Optional component  
Inventory rack (IR-220U)  
Temperature recorder + Mounting kit (MTR-G85, MTR-85H+MDF-S3085)  
Back-up system (CVK-UB2): LCO2  
Interface board (MTR-480), LAN interface board (MTR-L03)  
Data acquisition system (MTR-5000)  
Note:  
• Design or specifications will be subject to change without notice.  
*: Power source of the automatic temperature recorder shall be 220 V.  
- 71 -  
29  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SPECIFICATIONS  
Name  
Model  
Ultra-Low Temperature Freezer  
MDF-U7486S MDF-U7486SC  
External dimensions  
Internal dimensions  
Effective capacity  
Exterior  
W1130 x D875 x H1990 (mm)  
W870 x D600 x H1280 (mm)  
668 L  
Painted steel  
Interior  
Painted steel  
Outer door  
Inner door  
Painted steel  
ABS resin panel with stainless frame, 2 doors  
Shelf  
Stainless steel, 3 shelves (adjustable)  
W848 x D533 (mm), Load; 50 kg/shelf  
Access port  
17 mm diameter, 2 locations (back, bottom left corner)  
Insulation  
Compressor  
Evaporator  
Rigid polyurethane foamed-in place  
Hermetic type, Output; 1100 W (high stage side), 1100 W (low stage side)  
Tube on sheet type  
Condenser  
Refrigerant  
Temperature controller  
Temperature display  
Thermal sensor  
Alarm  
Fin and tube type (high stage side), Shell and tube type (low stage side)  
R-404A (high stage side), R-508B (low stage side)  
Microcomputer control system  
Digital display  
Platinum resistance (Pt 1000)  
High temp. alarm, Low temp. alarm, Power failure alarm  
Filter check, Battery check, Fan motor check  
Allowable contact capacity: 30 V DC, 2 A  
Remote alarm contact  
Battery  
Accessories  
Weight  
Nickel-metal-hydride battery, 6 V DC, 1100 mAh, Auto-recharge  
1 set of key, 1 scraper, 1 stick for air intake port cleaning  
360 kg  
None  
365 kg  
Built-in  
Voltage booster  
Optional component  
Inventory rack (IR-220U)  
Temperature recorder + Mounting kit (MTR-G85, MTR-85H+MDF-S3085)  
Back-up system (CVK-UB2): LCO2  
Interface board (MTR-480) , LAN interface board (MTR-L03)  
Data acquisition system (MTR-5000)  
Note:  
• Design or specifications will be subject to change without notice.  
*: Power source of the automatic temperature recorder shall be 220 V.  
- 72 -  
30  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PERFORMANCE  
Model  
MDF-U5486S  
MDF-U5486SC (for U.S.A. only)  
MDF-U7486S  
MDF-U7486SC (for U.S.A. only)  
Cooling performance  
Temperature control range  
Power source  
-86oC at the center of the chamber (ambient temperature; 30oC, no load)  
-50oC to -86oC (ambient temperature; 30oC, no load)  
220 V, 60 Hz  
Rated power consumption  
Noise level  
Maximum pressure  
1180 W  
51 dB [A] (background noise; 20 dB)  
2.7 MPa  
1210 W  
Note : The unit with CE mark complies with EC directives 89/336/EEC.  
- 73 -  
31  
Download from Www.Somanuals.com. All Manuals Search And Download.  
CAUTION  
Please fill in this form before servicing.  
Hand over this form to the service engineer to keep for his and your safety.  
Safety check sheet  
1. Freezer contents :  
Risk of infection:  
ʿYes  
ʿNo  
ʿYes  
ʿYes  
ʿYes  
ʿNo  
ʿNo  
ʿNo  
Risk of toxicity:  
Risk from radioactive sources:  
(List all potentially hazardous materials that have been stored in this unit.)  
Notes :  
2. Contamination of the unit  
Unit interior  
ʿYes  
ʿNo  
No contamination  
Decontaminated  
Contaminated  
Others:  
ʿYes  
ʿYes  
ʿYes  
ʿNo  
ʿNo  
ʿNo  
3. Instructions for safe repair/maintenance of the unit  
a) The unit is safe to work on  
ʿYes  
ʿYes  
ʿNo  
ʿNo  
b) There is some danger (see below)  
Procedure to be adhered to in order to reduce safety risk indicated in b) below.  
Date :  
Signature :  
Address, Division :  
Telephone :  
Product name:  
Model:  
MDF-  
Serial number:  
Date of installation:  
Ultra-low temperature  
freezer  
Please decontaminate the unit yourself before calling the service engineer.  
- 74 -  
32  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SANYO Electric Co., Ltd  
Printed in Japan  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Pyramid Technologies Time Clock 3000HD User Guide
RBH Sound Speaker SI 744 User Guide
Ricoh Network Card BB HPL10A R User Guide
Ryobi Grinder AG400 User Guide
Ryobi Saw P501 User Guide
Samsung Bluetooth Headset WEP350 User Guide
Samsung Car Stereo System 400UX 2 User Guide
Samsung DVD Player DVD F1080W User Guide
Samsung Printer 4050N User Guide
Senco Nail Gun AirFree 25 User Guide