CONTENTS  
PRECAUTIONS FOR SAFE OPERATION  
CAUTIONS FOR USAGE  
P. 2  
P. 6  
P. 6  
ENVIRONMENTAL CONDITIONS  
STERILIZER COMPONENTS  
INSTALLATION  
P. 7  
P. 10  
P. 11  
BEFORE COMMENCING OPERATION  
OPERATING INSTRUCTIONS  
ALARM & SAFETY FUNCTIONS  
P. 12  
P. 17  
ROUTINE MAINTENANCE  
TROUBLE SHOOTING  
DISPOSAL OF UNIT  
SPECIFICATIONS  
P. 18  
P. 19  
P. 19  
P. 20  
PERFORMANCE  
P. 20  
P. 21  
SAFETY CHECK SHEET  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PRECAUTIONSFORSAFEOPERATION  
WARNING  
Do not use the unit outdoors. Current leakage or electric shock may result if the unit is exposed to  
rain water.  
Only qualified engineers or service personnel should install the unit. The installation by  
unqualified personnel may cause electric shock or fire.  
Install the unit on a sturdy floor. If the floor is not strong enough or the installation site is not  
adequate, this may result in injury from the unit falling or tipping over.  
Never install the unit in a humid place or a place where it is likely to be splashed by water.  
Deterioration of the insulation may result which could cause current leakage or electric shock.  
Never install the unit under the water main or vapor pipe. Deterioration of the insulation may  
result which could cause current leakage or electric shock.  
( \ j Never install the unit in a flammable or volatile location. This may cause explosion or fire.  
Never install the unit where acid or corrosive gases are present as current leakage or electric  
shock may result due to corrosion.  
Use a dedicated power source as indicated on the rating label attached to the unit.  
Remove dust from the power supply plug before inserting in a power source. A dusty plug or  
improper insertion may pose a hazard.  
Use a power supply outlet with ground (earth) to prevent electric shock. If the power supply outlet  
is not grounded (earthed), it will be necessary to install a ground by qualified engineers.  
x ~\ Never ground the unit through a gas pipe, water main, telephone line or lightning rod. Such  
grounding may cause electric shock in the case of an incomplete circuit.  
Do not insert metal objects such as a pin or a wire into any vent, gap or any outlet for inner air  
0
circulation. This may cause electric shock or injury by accidental contact with moving parts.  
( \ j Never store volatile or flammable substances in this unit. This may cause explosion or fire.  
Use this unit in a safe area if using poisonous, harmful or radioactive substances. Improper  
use may be harmful to your health or the environment.  
Disconnect the power supply to the unit prior to any repair or maintenance in order to prevent  
electric shock or injury.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PRECAUTIONSFORSAFEOPERATION  
!\WARNING  
Ensure you do not inhale or consume medication or aerosols from around the unit at the time of  
maintenance. These may be harmful to your health.  
Never splash water directly onto the unit as this may cause electric shock or short circuit.  
Never disassemble, repair, or modify the unit yourself. Any such work carried out by an  
unauthorized person may result in fire or injury due to a malfunction.  
Disconnect the power supply plug if there is anything wrong with the unit. Continued abnormal  
operation may cause electric shock or fire.  
If the unit is to be stored unused in an unsupervised area for an extended period, ensure that children  
do not have access and that doors cannot be closed completely.  
The disposal of the unit should be undertaken by appropriate personnel. Remove doors to  
prevent accidents such as suffocation.  
Prepare a safety check sheet when you request any repair or maintenance for the safety of service  
personnel.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PRECAUTIONS FOR SAFE OPERATION  
CAUTION  
Select a level and sturdy floor for installation. This precaution will prevent the unit from tipping.  
Improper installation may result in water spillage or injury from the unit tipping over.  
Connect the unit to a power source as indicated on the rating label attached to the unit. Use of  
any other voltage or frequency other than that on the rating label may cause fire or electric shock.  
tfB Fix the shelves securely. Incomplete installation may cause injury or damage.  
When removing the plug from the power supply outlet, grip the power supply plug, not the cord.  
Pulling the cord may result in electric shock or fire by short circuit.  
Never damage or break the power supply plug or cord. Do not use the supply plug if its cord is  
loose. This may cause fire or electric shock.  
Do not touch any electrical parts such as the power supply plug or any switches with a wet hand.  
This may cause electric shock.  
Check the setting when resuming the operation after power failure or switch off. The stocked  
materials may be affected if the setting is changed.  
Do not put a container with water or heavy articles on the unit. It may cause injury if the articles  
fall. Current leakage or electric shock may result form deterioration of insulation by spilled water.  
Do not climb onto the unit and do not put articles on the unit. This may cause injury by tipping or  
0
damage to the unit.  
Never touch the inside chamber, door glass, exhaust vent or inside of the door when the unit is  
operating with high temperature setting as it may cause burn.  
Ml Hold the handle when closing the door. This will reduce the likelihood of a trapped finger.  
Never place your hand on the glass or use excessive force on the glass. Intentional force may  
break the glass resulting in injury from the broken glass.  
0
Do not lean on the door. This may cause injury, current leakage, or electric shock if the unit tips over  
or door becomes detached.  
Disconnect the power supply plug before moving the unit. Take care not to damage the power cord.  
A damaged cord may cause electric shock or fire.  
• В Be careful not to tip over the unit during movement to prevent damage or injury.  
^^ Disconnect the power plug when the unit is not used for long periods.  
( \ ) Do not put the packing plastic bag within reach of children as suffocation may result.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
CAUTIONS FOR USAGE  
1. This unit is not explode-proof construction. Never store the materials which cause flammable or volatile  
gas and never store the flammable or volatile liquid in the unit.  
2. The door glass, inside door panel, and exhaust vent are very hot when the unit is used with high  
temperature setting. Never touch such portion directly with the hands.  
3. Avoid to open the door when the chamber temperature is more than 100°C as it is dangerous. The  
door opening may cause ignition of flammable material or break of equipment resulting from sudden  
temperature change.  
4. Do not block the air intake vent on the chamber floor by the stocked articles. The blockage can cause  
unstable chamber temperature and shorten the heater life.  
5. When cleaning, do not use brushes, acids, benzine, thinner, soap, cleaner or hot water. These will  
cause discoloring or damage to coated surfaces.  
On plastic or rubber parts, they will cause  
transformation, discoloration or degeneration. Never apply volatile chemicals (like benzine etc.) on plastic  
or rubber parts. When neutral detergent is used, be sure to wipe it up thoroughly with a wet cloth  
afterwards. Before cleaning the box, refer to page 18.  
6. When placing objects inside, remove traces of water as much as possible.  
7. When stacking the unit, always use stacking kit available as an optional component.  
ENVIRONMENTAL CONDITIONS  
This equipment is designed to be safe under the following conditions (based on the IEC 1010-1):  
1.Indoor use;  
2. Altitude up to 2000 m;  
3. Ambient temperature 5°C to 35°C  
4. Maximum relative humidity 80% for temperature up to 31 °C decreasing linearly to 50% relative humidity  
at 40°C;  
5. Mains supply voltage fluctuations not to exceed ±10% of the nominal voltage;  
6. Other supply voltage fluctuations as stated by the manufacturer;  
7. Transient overvoltages according to Installation Categories (Overvoltage Categories) II; For mains  
supply the minimum and normal category is II;  
8. Pollution degree 2 in accordance with IEC 664.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STERILIZER COMPONENTS  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STERILIZER COMPONENTS  
1. Door: During use, inner panel of the door becomes extremely hot; Be careful.  
2. Observation window: During use, glass and its surrounding become very hot; Be careful.  
3. Leveling foot: Made of rubber. Can be adjusted by screws.  
4. Fan (inside): Propeller type. To cool down the heat discharge plate and fan motor.  
5. Control panel: Refer to page 9.  
6. Separation plate: Do not place objects directly on the plate. Especially if the holes are covered, the  
proper temperature cannot be obtained.  
7. Heater box (inside): The heater is attached under separation plate. The sirocco fan is also attached  
in this box.  
8. Packing: Be careful not to scratch.  
9. Key lock switch: Put the switch to "OFF" when you set the operating condition. After the setting, put  
the switch to "ON" to prevent the set condition from changing by accidental contact.  
10. Main power switch with circuit breaker: Main switch for all power. When the operation of the unit  
is stopped by this circuit breaker, contact with a dealer or a service station after disconnected the power  
supply plug.  
11. Exhaust air vents: The air inside the oven is exhausted from these vents. Be careful not to block  
these vents.  
12. Handle: Pull the knob of inside of the handle to open the door.  
13. Rack support: This can be adjusted to change the height.  
14. Rack: They can be slid forward.  
15. Exhaust air vent: The exhaust is controlled by opening/closing the vent. During use, the  
temperature of this section is extremely high; Be careful.  
16. Temperature sensor (installation position): Be careful not to allow objects to touch the sensor or  
scratch it.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8
STERILIZER COMPONENTS  
Control panel and keypad  
1
2
3
4
5
6
7
8
1. Power switch (POWER): Power switch of the control panel.  
2. Alarm buzzer stop key (BZ): Press this key to silence the buzzer in the event that the alarm operates  
and the buzzer sounds. Press it once again to reactive the buzzer.  
3. Timer mode select key (TIMER MODE): By pressing this key, the timer mode is selected. Refer to  
the "Timer function" on page 13.  
4. Digit shift key ( • • ) : Pressing this key in the setting mode causes the changeable digit to shift.  
5. Numerical value shift key (^ ): Pressing this key in the setting mode causes the numerical value to  
shift.  
6. Call key (CALL): By pressing this key, the unit enters the setting mode, and the digits that can be set  
flash, except that only set value display mode is available when unit is running.  
7. Delay timer key (DELAY): Pressing this key results in delayed starting of running.  
8. Run/Stop key (RUN/STOP): This key is for start/stop the running.  
9. Digital temperature/timer indicator (TEMP/TIME): This indicator shows the temperature or the time.  
10. Delay timer lamp: This lamp lights when the delay timer is active.  
11. Start lamp: This lamp lights when the unit is running.  
12. Alarm lamp: This lamp lights when the unit is warning condition.  
13. Set value lamp: This lamp lights when setting mode or set value display mode as the unit is running.  
14. Timer mode 1 lamp: This lamp lights when timer mode 1 is active.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
INSTALLATION  
Installation site  
To operate this unit properly and to obtain maximum performance, install the unit in a location with the  
following conditions:  
1. A location not subjected to direct sunlight or direct air flow from an air conditioner  
2. A location with adequate ventilation  
Leave at least 30 cm around the unit for ventilation. Poor ventilation will result in a reduction of the  
performance.  
3. A location away from heat generating sources  
Avoid installing the unit near heat-emitting appliances such as gas ranges or stoves.  
4. A location with a sturdy and level floor  
WARNING  
Install the unit on a sturdy floor. If the floor is not strong enough or the installation site is not adequate,  
this may result in injury from the unit falling or tipping over.  
Select a level and sturdy floor for installation. This precaution will prevent the unit from tipping.  
Improper installation may result in water spillage or injury from the unit tipping over.  
5. A location without flammable or corrosive gas  
WARNING  
Never install the unit in a flammable or volatile location. This may cause explosion or fire.  
Never install the unit where acid or corrosive gases are present as current leakage or electric shock  
may result due to corrosion.  
6. A location not prone to high humidity  
WARNING  
Do not use the unit outdoors. Current leakage or electric shock may result if the unit is exposed to rain  
water.  
Never install the unit in a humid place or a place where it is likely to be splashed by water.  
Deterioration of the insulation may result which could cause current leakage or electric shock.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
10  
INSTALLATION  
Installation  
1. Remove the packaging materials and tapes  
Remove all transportation packaging materials and tapes. Open the doors and ventilate the unit. If the  
outside panels are dirty, clean them with a neutral detergent and wipe it up with a wet cloth.  
2. Adjust the leveling feet  
Extend the leveling legs by rotating them counterclockwise so they contact the floor or bench. Ensure the  
unit is level.  
3. Ground (earth)  
WARNING  
Use a power supply outlet with ground (earth) to prevent electric shock. If the power supply outlet is  
not grounded, it is necessary to install a ground by qualified engineers.  
Never ground the unit through a gas pipe, water main, telephone line or lightning rod. Such  
grounding may cause electric shock in the case of an incomplete circuit.  
BEFORE COMMENCING OPERATION  
Prerunning  
When using the unit for the first time after purchasing, operate the unit without objects inside.  
1. Install the racks in the chamber.  
2. Set the temperature at 200°C and operate the unit for 20 minutes.  
3. Leave the unit as it is until the chamber temperature is cool enough.  
4. Ventilate the room when opening the chamber door as the smoke with a strong odor is exhausted.  
5. Keep the chamber door opened for a while until the odor is eliminated.  
Note:  
Some odor may be remained after prerunning. Such residue is eliminated gradually during usage.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
OPERATING INSTRUCTIONS  
Basic operating flow chart of control panel  
1. Start of operation when the input of setting is conducted.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
12  
OPERATING INSTRUCTIONS  
Timer function  
The timer function has two modes as follows:  
1. Normal mode: starts immediately after start of the running  
2. Timer mode 1: starts when the chamber temperature is 2.0°C lower than the set point.  
When conducting the sterilization, always select "Timer mode 1".  
On timer mode 1, the timer starts when the chamber temperature is 2.0°C lower than the set point. After  
timer starting, the temperature indicator blinks, alarm lamp lights, and intermittent tone sounds if the  
chamber temperature is 5.0°C or more lower than set point due to the door opening, etc.  
In spite of the alarm condition, the unit continues to run. The timer is reset to initial setting when the  
chamber temperature is 2.0°C lower than the set point again.  
Set point  
Start  
Normal mode  
Timer mode 1  
(For sterilization)  
Sterilization  
Refer to the table below when setting the sterilization temperature and duration.  
Sterilization temperature  
135to145°C  
Sterilization duration  
3 to 5 hours  
160to170°C  
2 to 4 hours  
180to200°C  
0.5 to 1 hour  
In case of sterilizing a lot of samples, extend the sterilization time for more than an hour.  
• When sterilizing the combustibles, such as cotton capped tube and instrument which are covered with  
paper, do not put them on the shelves directly. Use the wire basket and upper shelves instead.  
• When sterilizing the pipettes or the glass pipes, waterdrop must be dried out clearly, otherwise  
sterilization will not be completed.  
• Take out the articles in the chamber when the chamber temperature is lower than 50°C.  
Note:  
Do not open the door during sterilization.  
Do not open the door when the chamber temperature is higher than 100°C.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
OPERATING INSTRUCTIONS  
Stable running  
Table below shows the basic procedure for setting the chamber temperature. Perform key operations in  
the sequence indicated in the table. The example in the table is based on the assumption that the desired  
temperature is 50°C and continuous running.  
50°C  
Time  
Basic operation sequence (Example: Chamber temperature; 50 C, Continuous running)  
Indication after operation  
Key operated  
Description of operation  
О
О
DELAY  
START  
О ALARM  
О
0
SET VALUE  
MODE  
d Ь  
1
2
3
1
Press the power key.  
POWER  
(The current chamber temperature is displayed.)  
0
ALARM  
DELAY  
START  
SET VALUE  
О
(Delay time reset)  
DELAY  
0 0  
0 0  
О MODE 1  
Press the delay timer key.  
(The fourth digit of the indicator flashes.)  
О ALARM  
О
DELAY  
START  
SET VALUE  
Press the delay timer key to reset  
the delay time.  
DELAY  
DELAY  
О MODE 1  
0 0  
E5  
0 0  
О
О
DELAY  
START  
О ALARM  
SET VALUE  
Press the delay timer key and the  
current chamber temp, is displayed.  
О
4
О MODE 1  
О
О
DELAY  
START  
О ALARM  
SET VALUE  
(Temperature set)  
Press the call key.  
к
0
CALL  
(The third digit of the indicator 1lashes.)  
О ALARM  
О
О
DELAY  
START  
Set the temp, to 050 with the digit  
shift key and the numeric value  
shift key.  
0
SET VALUE  
MODE  
6
7
8
1
5 0  
0
О ALARM  
О
О
DELAY  
START  
(Time set)  
0
SET VALUE  
MODE  
a
о 0 U  
CALL  
1
Press the call key.  
(The fourth digit of the indicator flashes.)  
О ALARM  
О
О
DELAY  
START  
Set the time to --:- with the digit  
shift key and numeric value shift  
key.  
SET VALUE  
0 MODE  
1
_
-I  
- -  
0
0
ALARM  
SET VALUE  
О
DELAY  
RUN  
(Start running)  
START  
с
E
STOP  
О MODE 1  
9
Press the run/stop key.  
(The unit runs.)  
Note:  
When the delay time is not set, operation 2, 3, and 4 is not necessary.  
-:— of the indicator is displayed by the numerical value shift key at fourth digit flashing.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
OPERATING INSTRUCTIONS  
Stable running with delay time  
50°C  
Delay time  
Basic operation sequence (Example: Chamber temperature; 50°C, Delay time; 1 hour)  
Description of operation  
Key operated  
Indication after operation  
О ALARM  
О SET VALUE  
О MODE 1  
О
О
DELAY  
START  
POWER  
E5\  
Press the power key.  
(The current chamber temperature is displayed.)  
О ALARM  
DELAY  
(Delay time reset)  
• SET VALUE  
О MODE 1  
О
START  
DELAY  
00 00  
Press the delay timer key.  
(The fourth digit of the indicator flashes.)  
Set the time to 01:00 with the digit  
shift key and the numeric value  
shift key.  
О ALARM  
О
DELAY  
START  
SET VALUE  
••  
О MODE 1  
0 / 00  
E\5  
О ALARM  
О SET VALUE  
О
О
DELAY  
START  
Press the delay timer key and the  
current chamber temp, is displayed.  
DELAY  
CALL  
О MODE  
1
О ALARM  
SET VALUE  
О MODE  
О
О
DELAY  
START  
(Temperature set)  
Press the call key.  
0 DO  
1
(The third digit of the indicator flashes.)  
О ALARM  
О
О
DELAY  
START  
Set the temp, to 050 with the digit  
shift key and the numeric value  
shift key.  
SET VALUE  
••  
О MODE  
1
0 5 0  
О ALARM  
SET VALUE  
О MODE 1  
О DELAY  
START  
(Time set)  
О
0 0 DO  
CALL  
Press the call key.  
(The fourth digit of the indicator flashes.)  
О ALARM  
О DELAY  
Set the time to --:-- with the digit  
shift key and numeric value shift  
key.  
SET VALUE  
О
START  
О MODE  
1
••  
О ALARM  
О SET VALUE  
° MODE 1  
DELAY  
START  
(Start running)  
RUN  
m
STOP  
Press the run/stop key.  
(The unit runs.)  
Start of operation with no change of setting  
Key operated  
Indication after operation  
Description of operation  
О ALARM  
О SET VALUE  
О MODE 1  
О
О
DELAY  
START  
00 00  
Press the power key.  
POWER  
(The current chamber temp, is displayed.)  
О ALARM  
• SET VALUE  
О
О
DELAY  
START  
/ BID  
Press the call key  
CALL  
°
MODE  
1
(The third digit of the indicator flashes.)  
О ALARM  
SET VALUE « S T A R T  
О MODE 1  
О DELAY  
RUN  
О
5
E
Press the run/stop key.  
STOP  
(The unit runs.)  
Download from Www.Somanuals.com. All Manuals Search And Download.  
-IK  
OPERATING INSTRUCTIONS  
Sterilization with timer mode 1  
200°C  
2 hours  
lours)  
Basic operation sequence (Example: Chamber temperature; 180°C, Delay time; 2 I  
Key operated  
Indication after operation  
Description of operation  
0 ALARM  
SET VALUE  
О MODE  
О
О
DELAY  
START  
i
1
iE151  
0
Press the power key.  
1
POWER  
(The current chamber temperature is displayed.)  
О ALARM  
SET VALUE  
0 MODE  
О
DELAY  
START  
'
-
~В В ВD  
(Release of delay set time)  
Press the delay timer key.  
DELAY  
1
2
3
4
(The fourth digit of the indicator flashes.)  
О ALARM  
• SET VALUE  
О
DELAY  
START  
Press the delay timer key to reset  
the delay time.  
DELAY  
DELAY  
0 В В 0  
О MODE  
1
0 ALARM  
О SET VALUE  
О
О
DELAY  
START  
Press the delay timer key and the  
current chamber temp, is displayed.  
E 5  
0 MODE  
1
0 ALARM  
• SET VALUE  
О DELAY  
.
> В' *-, В D  
О
START  
(Temperature set)  
Press the call key.  
CALL  
lashes.)  
(The third digit of the indicator i  
О ALARM  
О
О
DELAY  
START  
1
Set the temp, to 180 with the digit  
shift key and the numeric value  
shift key.  
• SET VALUE  
О MODE 1  
6
7
••  
В п  
t
О ALARM  
• SET VALUE  
О
О
DELAY  
START  
(Time set)  
1
В
0 В  
CALL  
Press the call key.  
(The fourth digit of the indicator flashes.)  
0 ALARM  
О
0
DELAY  
START  
Set the time to 02:00 with the digit  
shift key and numeric value shift  
key.  
SET VALUE  
MODE  
8
9
••  
1
в E 0  
t
В
(Timer mode 1 set)  
TIMER  
MODE  
0 В  
E В  
Press the timer mode select key.  
(The timer mode 1 lamp is ighted.)  
О ALARM  
О SET VALUE  
MODE  
О
DELAY  
START  
RUN  
1
0
(Start of operation)  
q
E _7  
STOP  
1
Press the run/stop key.  
(The unit runs.)  
Download from Www.Somanuals.com. All Manuals Search And Download.  
ALARMS & SAFETY FUNCTIONS  
This unit has the alarms and safety functions shown in table below, and also self diagnostic functions.  
Alarms and safety functions  
Alarm &  
Buzzer  
Safety operation  
Indication  
Situation  
Safety  
If the internal temperature  
deviates from the set  
temperature by +10°C or more  
(and if timer mode 1, by -5°C or  
more)  
Automatic set  
temperature  
Alarm  
Alarm lamp lights.  
All digits on the  
indicator flash.  
Intermittent tone  
If a key operation is not  
performed for about 45 sec. in  
each setting mode.  
The setting mode  
is cancelled.  
Auto return  
Normal display mode.  
Key input is  
disabled.  
When the key lock switch is  
turned ON.  
Key lock switch  
Temperature  
sensor  
abnormality  
Alarm lamp lights.  
E01 is displayed on  
the indicator.  
If the temperature sensor goes  
open circuit.  
Intermittent tone Heater OFF  
Alarm lamp lights.  
E02 is displayed on  
the indicator.  
Intermittent tone  
Heater OFF  
If the triac goes open circuit.  
If the triac goes short circuit.  
If the relay goes short circuit.  
Triac  
abnormality  
Alarm lamp lights.  
E03 is displayed on  
the indicator.  
Intermittent tone Heater OFF  
Alarm lamp lights.  
E04 is displayed on  
the indicator.  
Heater OFF  
Intermittent tone  
Relay  
If the relay goes open circuit or  
fan motor and heater goes short  
circuit. Or the ambient temp,  
of circuit is 65°C or more.  
(when unit is not running)  
The security circuit is activated  
by independent temp, sensor if  
the chamber is abnormal over-  
heating.  
abnormality  
Alarm lamp lights.  
E05 is displayed on  
the indicator.  
Intermittent tone Heater OFF  
Heater OFF  
Without display  
change.  
forcedly by  
Continuous tone  
external circuit.  
Independent  
over-heat  
protection  
(When unit is running)  
Continuous tone  
(When the temp,  
s decreased,  
Alarm lamp lights.  
E05 is displayed on  
the indicator.  
Heater OFF  
forcedly by  
external circuit.  
The security circuit is activated  
by independent temp, sensor if  
the chamber is abnormal over-  
heating.  
ntermittent tone)  
* The buzzer tone resulting from the independent over-heat protection cannot be stopped with the BUZZER  
key. Turn off the main switch.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
ROUTINE MAINTENANCE  
LWARNING  
Always disconnect the power supply to the unit prior to any repair or maintenance of the unit in  
order to prevent electric shock or injury.  
Ensure you do not inhale or consume medication or aerosols from around the unit at the time of  
maintenance. These may be harmful to your health.  
CAUTION  
Always put on dry gloves to protect hands at the time of maintenance. Failure to use gloves may  
result in cuts or abrasions from any sharp edges or corners.  
Note:  
Never attempt to directly spray water on the heater box or the inside of the oven as it is very dangerous.  
In addition, never use volatile or combustible chemicals to clean the inside.  
Cleaning of unit  
Cleaning the inside  
• Remove all racks from the inside.  
• Clean the inside using a soft cloth damped with neutral detergent. Afterwards, wipe off with a cloth  
washed in clean water.  
• Remove the separation plate at the bottom of the oven and wipe off any particles in the heater box, using  
a soft cloth damped with water.  
Cleaning the frame  
• Clean the frame using a soft cloth damped with neutral detergent. Afterwards, wipe it up the detergent  
with a wet cloth.  
Cleaning the rack  
• To clean the rack, place it in a tub of warm water mixed with neutral detergent and wipe with a sponge or  
a soft cloth. Racks subject to high temperature will naturally become colored. This is a natural  
coloration, acknowledge this.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
18  
TROUBLE SHOOTING  
If the unit malfunctions, check out the following before calling for service.  
The unit does not operate at all  
1. The unit is not plugged correctly into a power outlet.  
2. The circuit breaker at the power source is active.  
3. A power failure has occurred.  
4. A fuse has blown.  
The key operation is disabled  
1. The key lock function is set in OFF mode.  
If the alarm function operates  
If the alarm function and the buzzer operates, check the cause using the following procedure.  
[At the beginning of operation]  
1. The chamber temperature is not equal to the set value.  
[During operation]  
1. The set temperature value was changed.  
2. The door was left open for a long period.  
3. A low temperature load was placed inside the unit.  
In these case, if the unit is left as it is, the alarm will eventually clear itself.  
If the chamber temperature is not equal to the set temperature  
1. The temperature in the vicinity is too high.  
The ambient temperature must always be at least 5°C less than the set temperature.  
2. The unit is installed tilted.  
DISPOSAL OF UNIT  
AWARNING  
If the unit is to be stored unused in an unsupervised area for an extended period ensure that children do  
not have access and doors cannot be closed completely.  
The disposal of the unit should be undertaken by appropriate personnel. Always remove doors to  
prevent accidents such as suffocation.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1Q  
SPECIFICATIONS  
Sterilizer  
Name  
Model  
MOV-112S  
MOV-212S  
W730 x D645 x H870 (mm)  
W600 x D500x H500 (mm)  
150 L  
External dimensions  
Internal dimensions  
Effective capacity  
Exterior  
W580 x D595 x H820 (mm)  
W450 x D450x H450 (mm)  
90 L  
Electrically galvanized steel plate, Acrylic resin painted  
Stainless steel plate (SUS 304)  
Electrically galvanized steel plate, Acrylic resin painted  
Reinforced glass x 3, W150 x H380 mm  
Glass wool  
Interior  
Door  
Observation window  
Insulation  
Rack  
Stainless steel plate, Stainless wire (SUS 304)  
Sensor K, PID control  
Temperature controller  
Temperature display  
Timer  
Digital display  
Electronic timer with delay timer  
15A  
Circuit breaker  
Overheat prevention  
mechanism  
Built-in thermister (electric circuit), Thermal guard  
Heater  
1.1 kW  
1.2 kW  
Fan motor  
Accessories  
Weight  
Output 10 W  
2 racks, 4 rack supports  
50 kg  
3 racks, 6 rack supports  
66 kg  
Note: Design or specifications will be subject to change without notice.  
PERFORMANCE  
Model  
MOV-112S  
MOV-212S  
Usable environment condition  
Extent chamber temperature  
Temperature control range  
Temperature; 0°C to 35°C  
Ambient temperature +5°C to 200°C  
±4°C  
Time of reaching maximum  
Temperature (0°C -» 200°C)  
60 minutes  
70 minutes  
AC110V 60Hz, 1030W  
AC115to120V 60Hz, 1130W  
AC220 50/60Hz, 1130W  
AC110 to 120V 60Hz, 1230W  
AC220V 50/60HZ, 1230W  
AC230 to 240V 50Hz, 1230W  
Power source, Rated power consumption  
AC230 to 240V 50Hz, 1130W  
Note: The unit with CE mark complies with EC directives 89/336/EEC, 93/68/EEC and 73/23/EEC.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
CAUTION  
Please fill in this form before servicing.  
Hand over this form to the service engineer to keep for his and your safety.  
Safety check sheet  
1. Refrigerator contents :  
•Yes  
•Yes  
•Yes  
•Yes  
No  
No  
No  
No  
Risk of infection:  
Risk of toxicity:  
Risk from radioactive sources:  
(List all potentially hazardous materials that have been stored in this unit.)  
Notes :  
2. Contamination of the unit  
Unit interior  
•Yes  
•Yes  
•Yes  
•Yes  
No  
No  
No  
No  
No contamination  
Decontaminated  
Contaminated  
Others:  
3. Instructions for safe repair/maintenance of the unit  
a) The unit is safe to work on  
DYes  
DYes  
DNo  
DNo  
b) There is some danger (see below)  
Procedure to be adhered to in order to reduce safety risk indicated in b) below.  
Date :  
Signature :  
Address, Division :  
Telephone :  
Serial number:  
Date of Installation :  
Product name :  
Sterilizer  
Model :  
MOV-112S, MOV-212S  
Please decontaminate the unit yourself before calling the service engineer.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
91  

PYLE Audio Speaker PDWT5 User Guide
RCA Car Video System DRC6379T User Guide
RCA Microwave Oven RMW742 User Guide
RCA Portable CD Player RP2910 User Guide
Reebok Treadmill RBTL078090 User Guide
Rolls Musical Instrument Rm203x User Guide
Rolls Stereo Receiver DB 25b User Guide
Ryobi Drill HD1800M User Guide
Sanyo Battery Charger N 1000SCR User Guide
Sanyo Flat Panel Television DP23845 User Guide