| 	
		 OPERATION OF THE PHILIPS CM-200 FEG-TEM   
					When not in use, the CM-200 should be in the MICROSCOPE ON configuration with the HIGH   
					TENSION ON (illuminates green when the high tension is on).   
					. The microscope is normally never turned off.   
					Preliminary   
					• 
					Add liquid Nitrogen to the Dewar next to the sample holder. (Anticontamination device:) Place the   
					metal-clad glass vacuum dewar into its holder to the right of the column so that the soft copper wire   
					"beard" hangs inside it. The glass viewing windows should be covered for safety. Fill the dewar   
					nearly full with liquid nitrogen and place the styrofoam cap on top.   
					• 
					• 
					Log in (in the hallway)   
					Go to vacuum page by pressing button Ready then button Vacuum. Make sure pressure reading of   
					IGP is below 10, preferably at 5 before proceeding to the next step.   
					1. Specimen Holder Removal, Loading, and Insertion   
					Never remove or insert the specimen holder when the red indicator light (on the front of the   
					compustage housing) is on. Do not touch the leading edge of the specimen holder (from the o-ring   
					to the tip) with ungloved hands. This portion resides inside the vacuum and must be kept clean.   
					There are several types of sample holders. The first type is single tilt (tilt in X or Alpha direction).   
					The second type is double tilt (tilt in both X and Y directions). Cold holder consists of a sample   
					holder and a chamber to contain liquid nitrogen. Thermometer and heating systems are attached.   
					Hearing holder can be used to hold samples that needed to be heated (up to 700°C) during   
					microstructure exploring.   
					1.1 REMOVAL: To remove the specimen holder from the column, carefully pull the round black   
					handle straight out until it stops, and hold it firmly so that it does not get pulled back into the 'scope   
					by the vacuum. Now rotate it clockwise until it stops again; it may now be pulled straight out   
					(carefully), free of the column. The specimen holder should be set down only on its Lucite stand.   
					1.2 SPECIMEN LOADING:   
					(i)   
					Use the pin tool (located in the Lucite stand, under the tip) to lift the grid clamping device at the   
					tip of the specimen holder).   
					(ii)   
					(iii)   
					Transfer your grid to the specimen holder using forceps. To make scanning the grid easier, you   
					may wish to orient one set of grid bars parallel to the long axis of the specimen holder.   
					Use the pin tool to carefully lower the clamping device onto the grid and lock it in place.   
					1.3 INSERTING THE SPECIMEN HOLDER INTO THE COMPUSTAGE:   
					(i)   
					(ii)   
					Touch the sample holder tip to one of the aperture handle to discharge the sample holder.   
					With the small pin in the holder tip at the 11:00 o'clock position, carefully insert the specimen   
					holder into the airlock entryway at the center of the compustage. Insertion of the holder will   
					initiate the pre-pumping sequence, and the red indicator light on the front of the compustage will   
					come on. Slide the holder in until it stops; at this point it will not go all the way in. (The data   
					1 
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				2.13 If the operator intends to expose TEM negatives using the built-in camera, it is best to ensure that   
					the CCD camera is retracted and shut off so that it cannot interfere with the shutter mechanism.   
					Using the Mac computer, whose CRT and keyboard rest on top on the CM-200 operating console,   
					the operator may open Digital Micrograph (if it is not already open), select MSC from the top   
					toolbar, and from that pull-down menu click to retract the CCD camera. The CCD camera control   
					unit, which rests to the left of the Mac CRT, may also be turned off using the red switch on the   
					front. Left.   
					3. Beam Alignment   
					These functions are performed without a specimen in the holder and may be performed without   
					inserting a specimen holder at all. If a holder is not in place, a radiation safety interlock prevents the   
					spot size from being changed to a number lower than 5 (the largest and brightest spot size is no. 1).   
					This will produce a much dimmer beam image than that seen with the holder in place.   
					3.1 Using the MAGNIFICATION knob, select 17,500x.   
					3.2 Defocus the electron beam by turning the INTENSITY knob clockwise to strongly overfocus the   
					C2 lens. Bring the beam back to crossover (the smallest image of the beam) and center it using the   
					SHIFT X/Y knobs.   
					3.3 Choose a C2 aperture (usually position 3, which is a 100-micron aperture) by rotating the largest   
					knurled knob on the topmost aperture control on the column to the desired position.   
					3.4 Using the INTENSITY knob, slowly overfocus (turn clockwise) the beam to coincide with the   
					diameter of the intermediate-sized black circle on the phosphorescent viewing screen. If the   
					defocused beam is not coincident with the circle, adjust it using the C2 aperture centering controls.   
					These are both the intermediate-sized knurled knob on the aperture rod and the knurled knob on the   
					side of the assembly (to the right). DO NOT TOUCH THE SMALLEST, INNERMOST KNOB   
					BECAUSE IT UNSCREWS THE APERTURE ROD. Repeat these steps (3 and 5) until the beam   
					spreads uniformly around the reference circle while being over focused (INTENSITY knob).   
					3.5 Using the INTENSITY knob, adjust the beam to crossover. Depress the FINE button (to the left of   
					the INTENSITY knob; the indicator light turns green when it is on) and use the INTENSITY knob   
					to sharpen the de-saturated filament image. Center the image using the SHIFT X/Y knobs.   
					3.6 Press the STIG button, on the right-hand instrument panel, to open the STIGMATOR CONTROL   
					page.   
					3.7 Press the COND key on the data monitor screen to select the C2 lens for astigmatism correction.   
					Use the MULTIFUNCTION X/Y knobs, while rotating the beam through crossover using the   
					INTENSITY knob, to obtain the cleanest, roundest de-saturated filament image possible.   
					3.8 Press the STIG button to return to the CONFIGURATION page.   
					3.9 Turn the FILAMENT knob clockwise to fully saturate the filament. When the saturation point (as   
					defined by the FIL LIMIT) has been reached, the microscope will emit a 'beep.'   
					3.10 Press the READY button and then the TEM key to return to TEM BRIGHTFIELD.   
					3.11 Using the lever to the left of the viewing chamber, lower the small phosphorescent screen into   
					place. Insert the beam stop (at the right-hand side top of the viewing chamber) so that it will be   
					visible across the small screen. Use the binoculars to observe the beam stop. To adjust the   
					binoculars for your eyes, first adjust the interpupillary distance so that you can see through them   
					with both eyes; then adjust each eyepiece so that the rough edges of the beam stop are in focus for   
					each eye. When you are done, retract the beam stop and lift the small screen back out of view.   
					4. Eucentric Height Adjustment   
					4.1   
					Set the MAGNIFICATION to 5,800x and use the X/Y JOYSTICK to center a small notable   
					feature of your specimen.   
					4.2   
					Focus your chosen feature using the knobs marked STEP SIZE. The outer knob changes the   
					focus; the inner knob (the step size adjustment) modifies the amount of focus change per 'click' of   
					the outer knob.   
					3 
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				4.3   
					From the TEM BRIGHTFIELD page, press COMPUSTAGE once; the COMPUSTAGE   
					REGISTER CONTROL page will appear.   
					4.4   
					4.5   
					Press A-WOBBLER; this will initiate back-and-forth tilting of the goniometer.   
					Use the Z control lever on the JOYSTICK to move the specimen up or down and thus minimize   
					apparent movement of the centered feature.   
					4.6   
					5.   
					When the feature moves only minimally or not at all, press the A-WOBBLER key to inactivate   
					tilting; then press the READY button to return to TEM BRIGHTFIELD.   
					Pivot Point Alignment   
					Ensure that the specimen is eucentric before performing this procedure. For this procedure you   
					may work with the OBJECTIVE APERTURE in place to protect your specimen.   
					5.1   
					5.2   
					Center (X/Y JOYSTICK) and focus (concentric knobs under STEP SIZE) an image feature at   
					24,500x (MAG knob).   
					Press the ALGN button to access the ALIGNMENT SELECTION page. (Note that the   
					alignments are divided into PROCEDURES (on the left side of the page) and DIRECT   
					alignments (most of which are on the right side of the page). Most users will not want to access   
					the PROCEDURES, which are long and complicated. For this set of instructions we will be using   
					only the DIRECT alignments.)   
					5.3   
					5.4   
					5.5   
					Using the INTENSITY knob, adjust the beam to crossover.   
					Press the beamcoils PIVOT POINT X key, on the right side of the page, so that it is highlighted.   
					Using the MULTIFUNCTION X/Y knobs, bring the two beam spots (the pivot points, on the   
					fluorescent viewing screen) together so that they overlap.   
					5.6   
					5.7   
					5.8   
					Center the coinciding spots using the SHIFT X/Y knobs.   
					Press the beamcoils PIVOT POINT Y key.   
					Using the MULTIFUNCTION X/Y knobs, bring the two beam spots together so that they   
					overlap.   
					5.9   
					Center the coinciding spots using the SHIFT X/Y knobs.   
					5.10   
					Press the ALGN button to exit the ALIGNMENT SELECTION page.   
					6.   
					Rotation Center Alignment   
					This procedure may be performed with the OBJECTIVE APERTURE in place to protect the specimen   
					and add contrast to the image.   
					6.1   
					6.2   
					Focus and center a feature of the specimen at 100,000x.   
					Press ALGN to open the ALIGNMENT SELECTION page. On the upper right-hand side of the   
					page, press ROT CENTER so that it becomes highlighted. Now either voltage or current   
					centering may be performed.   
					6.3   
					(i)   
					VOLTAGE CENTERING   
					On the lower right-hand side of the page, select VOLT (under rot center VOLT CURR) so that it   
					becomes highlighted. This will cause the high tension to modulate (the inner STEP SIZE knob   
					adjusts the amplitude of modulation). If the chosen feature shifts off center laterally, the beam is   
					not aligned along the optical axis of the microscope and must be corrected.   
					(ii)   
					Use the MULTIFUNCTION X/Y knobs to stabilize the feature at the center of the screen,   
					eliminating all lateral movement. The feature should appear to be pulsating.   
					(iii)   
					Press the ALGN button to return to the TEM BRIGHTFIELD page.   
					6.3.1 Current centering:   
					4 
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				(i)   
					On the lower right-hand side of the page, select CURR (under rot center VOLT CURR) so that it   
					becomes highlighted. This will cause the objective lens current to modulate (the inner STEP SIZE   
					knob adjusts the amplitude of modulation). If the chosen feature shifts off center laterally, the   
					beam is not aligned along the optical axis of the microscope and must be corrected.   
					Use the MULTIFUNCTION X/Y knobs to stabilize the feature at the center of the screen,   
					eliminating all lateral movement. The feature should appear to be pulsating.   
					(ii)   
					(iii)   
					Press the ALGN button to return to the TEM BRIGHTFIELD page.   
					7. Centering the Objective Lens Aperture   
					For this procedure a specimen must be in place. Choose an area for which it is acceptable to   
					sustain beam damage. With the OBJECTIVE APERTURE out, set the magnification to 5800x   
					and overfocus the beam by turning the INTENSITY knob clockwise from crossover.   
					7.1   
					Set the focus step size (inner STEP SIZE knob) to 5 and press the Zoom button to put the 'scope   
					in diffraction mode.   
					7.2   
					7.3   
					7.4   
					If necessary, adjust the CAMERA LENGTH to 620 mm using the MAG knob.   
					Center the diffraction spot using the MULTIFUNCTION X/Y knobs.   
					Using the focus (outer STEP SIZE) knob, refocus the beam to the smallest, brightest spot. It may   
					also be necessary to adjust the INTENSITY knob.   
					7.5   
					7.6   
					Insert the OBJECTIVE APERTURE by rotating the aperture displacement lever below it to the   
					left.   
					Apertures may be selected by rotating the largest knurled knob on the objective aperture assembly   
					to any one of four numbered positions. (The APERTURE MEMO lists the diameters and   
					positions of the apertures currently installed in the 'scope. It may be accessed from the TEM   
					BRIGHTFIELD page by clicking MODES and then CONFIGURATION.)   
					7.6.1 Once the aperture has been selected, center it using both the smaller knurled knob in the series on   
					the aperture assembly and the small knurled knob to the right. Remember not to manipulate the   
					smallest, innermost knob, which is not knurled and will unscrew the aperture rod.   
					7.6.2 Press Zoom to exit diffraction mode.   
					8. Objective Lens Astigmatism Correction   
					8.1   
					Select an area that may be imaged at high magnification without harming any desirable portions   
					of the specimen. Increase the magnification to 175,000x or higher and adjust the illlumination so   
					that the substructure or background grain of the specimen may be observed easily. The   
					INTENSITY will have to be modified and the beam will have to be recentered using the SHIFT   
					X/Y knobs (Deflectors) as the magnification is increased. This latter function may alternatively   
					be controlled using the RST button, on the panel to the left of the column.   
					Set the focus step size to 2 and obtain a slightly underfocused image for maximum contrast.   
					Press the STIG button to open the STIGMATOR CONTROL page. If it is not already   
					highlighted, press the OBJ key on this page.   
					8.2   
					8.3   
					8.4   
					8.5   
					Use the MULTIFUNCTION X/Y knobs, one at a time, to obtain the sharpest possible image of   
					the grain substructure.   
					Confirm that any astigmatism has been corrected by varying the focus (back and forth through   
					focus, from underfocus to overfocus) and watching to see if a "streaking" pattern emerges and   
					changes direction between under- and overfocus. If the astigmatism has been corrected, the   
					specimen will vary only in focus, with no pattern evident. Repeat steps 4 and 5 until no pattern is   
					apparent.   
					8.6   
					Press the STIG button again to return to the TEM BRIGHTFIELD page.   
					9. Condense Lens Astigmatism Correction   
					This procedure is necessary if the beam spot is not a symmetric circle when the intensity knob is   
					rotated.   
					5 
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				9.1   
					9.2   
					9.3   
					Press STIG button to open the STIGMATOR CONTROL page.   
					Select Condense Aperture.   
					Use the MULTIFUNCTION X/Y knobs, one at a time, such that the beam spot circle has a fixed   
					center when the beam intensity is changed.   
					10. High Resolution Images   
					10.1   
					10.2   
					10.3   
					Bring in the small screen into viewing chamber.   
					TV on.   
					Amplifier on.   
					11. Load Films   
					11.1   
					11.2   
					11.3   
					11.4   
					Press CAMERA AIR.   
					Put film in.   
					PUMP till to IGP less than 10.   
					Note films, along with film box should be put into a vacuum for more than 24 hours to degas.   
					Note: reset the negative numbers   
					6 
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 |