Fully automatic blood pressure monitor
705IT Int-0-0
1
Download from Www.Somanuals.com. All Manuals Search And Download.
English
OMRON 705IT is a compact, fully automatic blood
pressure monitor, operating on the oscillometric
principle. It measures your blood pressure and pulse
rate simply and quickly. It contains an intelligent “fuzzy
logic” system for controlled inflation,
known as “Intellisense”. This is an
advanced method of oscillometric
measurement and does not require pressure pre-
setting or re-inflation. OMRON is the world’s leading
manufacturer of blood pressure monitors with the goal
to always fulfill the need for reliable regular monitoring.
Slide the battery cover off in the direction of the
arrow.
Your blood
pressure
monitor:
Battery
installation /
replacement
Install or replace 4 “LR6” batteries so the +
(positive) and – (negative) polarities match the
polarities of the battery compartment as indicated.
the benefits
Replace the battery cover.
Set Button
Heart Symbol
Adjust Button
NOTES :
If the Low Battery Indicator (
) appears on the
AC Adapter Jack
USB Connector
display, replace all four batteries. Long-life alkaline
batteries are recommended.
Systolic Blood
Pressure
Remove the batteries if the monitor will not be
used for an extended period of time.
Diastolic Blood
Pressure
ON/OFF (O/I) button
START ( ) button
Replace batteries within 30 seconds. If the
batteries remain removed longer than that time,
the set date and time will be deleted.
Air Jack
Pulse Rate
Date/Time Display
Memory recall Button
Cuff (Applicable cuff for Adults 22 to 32 cm.)
Air Plug (Connects to Air Jack)
Green Mark
Air Tube
CD-ROM
USB Cable
4
5
Download from Www.Somanuals.com. All Manuals Search And Download.
English
• You can wrap the cuff either on your right or left arm.
When the batteries are installed, the display will
show 0:00.
How to set time
and date
How to apply
the cuff
(when wrapping on your left arm)
Insert air tube to air jack (on the left side of the
divice). The cuff must be airless.
Press and hold the SET button.
The year digits (2002) will flash.
Remove tight-fitting clothing from your upper arm.
Press the ADJUST button to advance the digit(s)
one at a time.
If you hold down the ADJUST button, the digits will
advance rapidly.
DATE/TIME SET
Sit in a chair with your feet flat on the floor and
place your arm on a table so that the cuff is at the
same level as your heart.
SET
ADJUST
Press SET button when desired number is on the
display to lock setting.
Put your arm through the cuff loop. The bottom of
the cuff should be approximately 1 to 2 cm above
the elbow. The green marker on the cuff should lie
over the brachial artery on the inside of the arm.
The tube should run down centre of arm
Repeat Step 3-4 for month and date.
approximately even with middle finger.
Bottom Edge 1 to 2 cm
Repeat Step 3-4 for hour and minutes.
Cuff Tube
Green Mark
6
7
Download from Www.Somanuals.com. All Manuals Search And Download.
English
Pull the cuff so that the top and bottom edges are
tightened evenly around your arm.
Press the (O/I) button.
How to take a
reading
a) All display symbols appear for approximately
one second.
b) When the monitor becomes ready to measure,
When the cuff is positioned correctly, close the
velcro fastener FIRMLY.
the (
) symbol appears on the display.
Press the (START ) button and remain still.
Make certain the cuff fits snugly around your arm.
The cuff should make good contact with your skin.
You should be able to fit your index finger
between the cuff and your arm easily, so you
can pull the cuff off and on.
As the cuff begins to inflate, the monitor
automatically determines your ideal inflation level.
Because this monitor detects the pulse even
during inflation, do not move your arm but remain
still until the entire measurement completes.
• If you want to stop the inflation or measurement,
press the (O/I) button. The monitor will stop
inflating and start deflating rapidly, then the
monitor will turn off.
Relax your arm and turn your palm upward.
Be sure there are no kinks in the air tubing.
Inflation
Inflation stops automatically and measurement is
started.
As the cuff slowly deflates, decreasing numbers
appear on the display and the (
) symbol
Deflation
flashes at every heartbeat. In rare circumstances,
a higher inflation may be necessary. In those
cases, the monitor reinflates the cuff up to 30
mmHg higher than initial inflation and restarts the
measurement.
Right hand measurement
8
9
Download from Www.Somanuals.com. All Manuals Search And Download.
English
This monitor has a memory capable of storing 28
sets of readings. Every time you complete the
measurement, the monitor automatically stores blood
pressure and pulse rate.
When the measurement is complete, the cuff
completely deflates and your blood pressure and
pulse rate are displayed.
How to use the
memory
function
NOTE :
The monitor automatically stores blood pressure
• Turn on the monitor by pressing the (O/I) button.
and pulse rate into the memory.
• Press the (
while the (
) button to recall stored readings
) symbol is displayed.
• When 28 sets of readings are stored in memory,
the oldest set will be deleted to store a new set.
Press the (O/I) button to turn the monitor off.
NOTE :
If you forget to turn the monitor off, it will
automatically shut itself off after five minutes.
Date and Time of stored readings
will be alternately displayed.
Instructions for special conditions:
• If your systolic pressure is known to be more than
220 mmHg, press and hold the (START ) button
until the monitor inflates 30 to 40 mmHg higher
than your suspected systolic pressure.
• To delete all the stored readings, press the (
and the (START ) button simultaneously.
) button
START
NOTE :
You can not delete the stored readings partially.
NOTE :
Do not apply more pressure than necessary.
The monitor will not inflate above 300 mmHg.
Continue to press button
START
Press these two buttons
at the same time.
10
11
Download from Www.Somanuals.com. All Manuals Search And Download.
English
How to uninstall the software
How to use the computer interface
1) Open the control panel
• Click the “Start” button
This blood pressure monitor has a computer interface and is supplied with software to
transfer the memory of the blood pressure monitor to your personal computer.
• Choose “Settings”
• Click “Control Panel”
Required operating environment
The following operating environment is required to operate the communication software as
supplied on the CD-ROM.
2) Open the “Add/Remove Prograns” icon by double clicking
3) Delete the application
• Click “BPFile” from the list
• Click the Add/Remove button
• .....The confirmation screen is displayed
• Click “Yes”
• .....”BPFile” is deleted
• Click “OK”
Personal computer
:
PC/AT compatible with CPU of 80486 or above or with
any Pentium processor.
Windows 98SE/ME/XP/2000
16 MB or more
Free area of 10 MB or more is necessary
640 x 480 pixels or more
CD-ROM drive
OS
:
:
:
:
:
:
Memory (RAM)
Hard disk
Display
Disk drive
Connection port
How to connect your 705IT blood pressure monitor to your PC
Free USB connection port
Connect your 705IT blood pressure monitor with your PC using the included USB cable.
The mini USB connector (small connector) should be connected to your 705IT blood
pressure monitor and the normal USB connector (large connector) should be plugged into
your personal computer.
How to install the software
1) Turn on your personal computer
Please ensure you connect the USB cable the correct way to your 705IT blood pressure
monitor, with the [O] symbol facing upwards.
......Your Windows™ desktop is displayed
2) Insert the CD-ROM in the disk drive
3) Open the control panel
• Click the “Start” button.
• Point the mouse pointer to “Settings” on the menu.
• Click “Control Panel”
4) Open the “Add/Remove Programs” by double clicking the icon
5) Execute the installation
• Click the “install button”
• Follow the instructions
• The software is installed on your personal computer
12
13
Download from Www.Somanuals.com. All Manuals Search And Download.
English
How to use the software
How to set the time
• Press the “TIME” button on your screen.
Open the BPFile application from the program list or your desktop and the following screen
is shown on the screen.
...The time and date are read from your 705IT blood pressure monitor
• Change the time and date if applicable in the white field
•
•
•
•
By pressing the “DATA” button you are able to transfer the memory of your 705IT blood
pressure monitor to your personal computer.
By pressing the “TIME” button you are able to set the date and time in your 705IT blood
pressure monitor.
By pressing the “ID” button you are able to set your personal name in your 705IT blood
pressure monitor.
By pressing the “CLOSE” button you can close the program.
How to set a patient ID in your 705IT blood pressure monitor
• Press the “ID” button on your screen
...The current patient ID is shown
• Change the patient ID if applicable in the white field
How to create a data file
• Press the “DATA” button on your screen
• Press the “Dir” button on your screen and
name the file.
• Press “OK”
• The file can be opened in most windows
programs like Excel, Word
14
15
Download from Www.Somanuals.com. All Manuals Search And Download.
English
• Before measuring, stretch your back and sit up
straight, breath normally 5-6 times, relax your
shoulders, arms and entire body.
Hints on taking
blood pressure
readings
Failure, causes and rectification
Failure and their possible causes
Rectification
• Slightly bend and support your elbow (e.g. on a
table).
E
Incorrect measurement is indicated by
• Do not move, talk or touch the device during
measurement.
1. A correct reading could not be obtained
because measurement was disturbed by
movement of the body.
1. Repeat the measurement keeping
perfectly still. Do not move your arm and
do not speak.
• Stress influences your blood pressure, so try to
relax before taking a measurement.
• To compare results, measurements must always
be taken from the same arm.
2. The cuff is not fitted correctly.
2. Check that the cuff is correctly fitted.
Then repeat the measurement.
• Do not place the cuff over thick clothes and do not
roll up your sleeve if it is too tight.
3. Your clothing has constricted your
blood flow.
3. Remove the item of clothing which
caused the constriction.
• Align the cuff with the height of your heart (breast
height). If necessary use a pillow or cushion.
4. There is still air in the cuff when the
monitor is switched off.
4. The unit may be defective. Please have
it examined and, if necessary, repaired
by OMRON.
• Coldness raises your blood pressure. It is
recommended to measure at room temperature
(+20°C).
The (START ) button was pressed
Wait for the ( ) symbol indicating
readiness before you press the (START
button.
• Relax before measurement. If necessary go to the
bathroom before measurement.
before the (
) symbol was displayed.
)
• Do not measure directly after bathing.
The blood pressure values displayed are
extremely low or high, or they are
implausible.
Refer to the instructions and then repeat
the measurement.
• If you wish to interrupt measurement prematurely,
press the (O/I) button. Inflation is stopped and the
air is automatically released from the cuff.
The display does not light up when the
(START ) button is pressed.
• Blood pressure measurements taken in a moving
vehicle may be influenced by vibration.
E
• If an error has occurred during measurement
displayed. Please refer to the next chapter.
is
1. The batteries are exhausted.
1. Check the batteries and, if necessary, fit
4 new batteries.
• The monitor should be switched off after
2. The +/- poles of the batteries have been
reversed.
2. Reinsert the batteries with the +/- poles
the right way round.
measurement to avoid running down the batteries
unnecessarily. If you forget to do so, the monitor
switches itself off automatically after 5 minutes.
3. The battery contacts are dirty.
3. Clean the battery contacts with a dry
cloth.
The batteries are weak or exhausted.
Fit new batteries.
The cuff pressure does not rise
although the pump motor can be heard.
Check that the air tube is properly
connected to the monitor. Push the tube
connector firmly into the socket.
problem with mem. function
Contact your Omron retailer or distributor.
Er
16
17
Download from Www.Somanuals.com. All Manuals Search And Download.
English
• For cleaning the monitor only use a soft, lightly
moistened cloth.
Do not use petroleum spirits, thinners or
similar solvents !
This device fulfils the provisions of the EC
Maintenance
and storage
Technical data
directive 93/42/EEC (Medical Device Directive).
This blood pressure monitor is designed
according to the European standard EN1060,
Non-invasive sphygmomanometers Part 1:
General requirements and Part 3: Supplementary
requirements for electromechanical blood
pressure measuring systems.
• Stains on the cuff can be carefully removed using
a moist cloth and soapy water.
Do not wash the cuff or make it wet !
• Keep your blood pressure monitor in the storage
case to protect it from dust and moisture.
Model:
OMRON 705IT
Display:
LCD digital display
Blood Pressure
Measurement Range:
Accuracy:
Pressure: 0 mm Hg to 299 mm Hg
Pulse: 40 – 180/min.
• Protect your monitor against vibrations and do not
leave it in a place where temperatures are
extremely low (below -20°C) or extremely high
(above 60°C) or with high humidity (+ 85%).
Pressure: ±3 mm Hg
Pulse: ±5% of reading
Memory:
28 measurements
• Do not use your monitor at very low temperatures
(below 10°C) or very high temperatures (above
40°C).
Inflation:
Fuzzy-Logic controlled by electric pump
Automatic pressure release valve
Capacitive pressure sensor
Oscillometric method
Deflation:
Pressure Detection:
Measurement Method:
Power Source:
• Do not fold the cuff or air tube together too tightly
and do not crease them.
4 Alkaline batteries 1.5V (Type LR6) or
AC/DC adapter (optional, 6V = 4W)
Battery Life:
Capacity of new batteries is approx. 300
measurements
• Do not carry out repairs of any kind yourself.
If a defect occurs, consult the OMRON distributor
or Customer Services or seek the advice of your
surgical appliance stockist or pharmacist.
Operating Temperature:
Storage Temperature:
Humidity:
10°C to 40°C
-20°C to 60°C
15 to 90% RH maximum
• The accuracy of this blood pressure monitor is
designed to last a long time.
Storage humidity:
Console Weight:
Outer Dimensions:
Cuff Dimensions:
10 to 95% RH maximum
Approximately 380g without batteries
Approximately 115 (l) mm x 177 (w) mm x 71 (h) mm
• In order to increase the lifetime of the batteries
take them out when you store the device for a
longer period than 2 months.
Approximately 140 mm x 480 mm
(arm circumference 22 – 32 cm)
Accessories:
Medium cuff, CD-ROM, USB cable, instruction
manual, storage case, battery set
• Electro-magnetic fields can influence the
measurement. Do not use a portable phone near
this unit.
NOTE: Subject to technical modification without prior notice
• Disposal of this product and used batteries should be carried out in accordance
with the national regulations for the disposal of electronic products.
= Type B
18
19
Download from Www.Somanuals.com. All Manuals Search And Download.
English
Calibration
OMRON spare
parts
Some useful information about blood pressure
What is Blood Pressure ?
The accuracy of this blood pressure monitor has
been carefully tested and is designed for a long
service life. It is generally recommended to have the
monitor inspected every two years to ensure correct
functioning and accuracy. Please consult your
authorised dealer or the OMRON Customer Service
at the address given on the packaging/attached
literature.
Blood pressure is a measure of the force of blood flowing against the walls of the
arteries. Arterial blood pressure is constantly changing during the course of the heart’s
cycle. The highest pressure in the cycle is called the SYSTOLIC BLOOD PRESSURE;
the lowest is the DIASTOLIC BLOOD PRESSURE. Both pressure readings, the
SYSTOLIC and DIASTOLIC, are necessary to enable a doctor to evaluate the status of
a patient’s blood pressure.
OMRON devices are especially designed for regular
blood pressure monitoring. OMRON therefore has
direct contact with medical specialists and takes
advice, on the design and features of its blood
pressure monitors.
Printer
Many factors such as physical activity, anxiety, or the time of day, can influence
your blood pressure. Blood pressure is typically low in the morning and increases from
afternoon to evening. It is lower in the summer and higher in the winter.
Blood pressure is measured in millimetres of mercury (mmHg) and measurements
are written with the systolic pressure before the diastolic e.g. A blood pressure written
as 140/90, is referred to as 140 over 90 mmHg.
Spare parts & accessories
AC adapter
Due to high company quality standards, OMRON
considers the main unit as a non-serviceable part
because of the necessity of proper calibration after
replacement of high-tech components.
High Blood Pressure
The discovery that you have high blood pressure is more often than not a chance
finding. People with high blood pressure usually feel well unless the blood pressure has
been high for some time, and complications have occurred. Such complications may
effect the heart, brain and other important organs. They may bring on ill health and
affect your enjoyment of life. It is because of the complications, rather than the high
blood pressure itself, that this condition is taken so seriously today.
Small cuff
Arm circumference 17 – 22 cm
High blood pressure is sometimes referred to as raised blood pressure, elevated blood
pressure, or hypertension. Most people with this condition require regular long-term
drug treatment, usually in the form of tablets. Some can lower their blood pressure
sufficiently by other means, such as changing diet and life-style.
People who have high blood pressure, and who also understand their condition, have
an advantage. They are more likely to attend for regular check-ups, to cooperate with
advised diets and changes in lifestyle, and to take medicines correctly. This
immediately leads to better health. It is hoped that, by a better understanding of high
blood pressure and how it is treated, you will take an active and informed part in your
health care, and so enjoy a better quality of life.
Medium cuff
Arm circumference 22 – 32 cm
Extra large cuff
Arm circumference 32 – 42 cm
20
21
Download from Www.Somanuals.com. All Manuals Search And Download.
What Causes High Blood Pressure ?
Blood pressure is only classed as high if it doesn’t go down when you rest. Permanently raised blood
pressure can be caused by several factors such as hardening of the arteries, smoking or drinking too
much alcohol. It can also be caused by cholesterol, a type of fat, building up inside your blood vessels
and making them narrower. Being under stress can add to the problem.
High blood pressure is also known to run in families hence if you find you have a high blood pressure it
is a good idea to test other closely related members of your family.
Fluctuation within a Day
(Example: 35-year-old male)
Blood pressure
(mmHg)
Breakfast
Rest room
Dinner
Sleep
Lunch
Systolic
blood
pressure
Diastolic
blood
pressure
Time
Brachial blood pressure data
Why is it a Good Thing To Measure Blood Pressure at Home ?
Having your blood pressure measured by a doctor can cause anxiety which is a cause itself of high
blood pressure. As a variety of conditions affect blood pressure, a single measurement may not be
sufficient for an accurate diagnosis.
The blood pressure measured first thing in the morning after getting up, before eating and while at rest,
is known as the fundamental blood pressure. In practice it is rather difficult to record the fundamental
blood pressure, but it is important to get readings in an environment that is as close as possible to this.
Classification of Blood Pressure by the World Heath
Organization
180
The World Health Organization (WHO) and the International
Society of Hypertention (ISH) developed the Blood Pressure
160
Classification shown in this figure.
140
This classification is based on the blood presure values
130
measured on people in a sitting position in outpatient
120
departments of hospitals.
*There is no universally accepted definition of hypotension. However, those
having the systolic pressure below 100 mmHg are assumed as hypotensive.