Omega Engineering Thermometer HH802U User Manual

Temperature Coefficient:  
INTRODUCTION  
0.1 times the applicable accuracy specification per °C from 0°C to 18°C and  
28°C to 50°C (32°F to 64°F and 82°F to122°F).  
Input Protection:  
24V dc or 24V ac rms maximum input voltage on any combination of input  
pins.  
Maximum Differential Common Mode Voltage (Maximum Voltage be-  
tween T1 and T2 during measurement): 1volt.  
Reading Rate: 1 sample/second.  
This instrument is a 5 digit, compact-sized portable digital thermometer de-  
signed to use external K type and J type thermocouple sensor. Temperature  
indication follows Reference Temperature/Voltage Tables (N.I.S.T. Mono-  
graph 175 Revised to ITS-90) for K type and J type thermocouples. Two K  
type thermocouple are supplied with the thermometer.  
SAFETY INFORMATION  
OPERATING INSTRUCTIONS  
It is recommended that you read the safety and operation instructions before  
using the thermometer.  
OPERATING INSTRUCTIONS  
MODEL HH802U  
DUAL INPUT K/J  
WARNING  
1.  
Power Switch  
To avoid electrical shock, do not use this instrument when working voltages  
at the measurement surface over 24V AC or DC.  
The  
key turns the thermometer on or off. In the SET mode the unit can-  
not be powered off. Exit SET mode to power off.  
DIGITAL THERMOMETER  
APO function mode  
WARNING  
To avoid damage or burns, do not make temperature measurement in mi-  
crowave ovens.  
Press  
power key for more than 6 seconds to disable the auto-power  
function. The display will show “APO OFF”.  
2. °C/°F Selecting the Temperature Scale  
Readings are displayed in either degrees Celsius(°C) or degrees Fahren-  
heit(°F). When the thermometer is turned on, it is set to the temperature  
scale that was in use when the thermometer was last turned off. To change  
the temperature scale, press the °C/°F key.  
Button (only Main display)  
Press this key to enter the Data Hold mode, the “HOLD” annunciator is dis-  
played. When HOLD mode is selected, the thermometer holds the present  
CAUTION  
Repeated sharp flexing can break the thermocouple leads. To prolong lead  
life, avoid sharp bends in the leads, especially near the connector.  
Input Connector:  
HH802U  
3.  
Accepts standard miniature thermocouple connectors (flat blades spaced  
7.9mm, center to center).  
readings and stops all further measurements. Press the  
key again to  
ENVIRONMENTAL  
0°C to 50°C (32°F to 122°F) <70% R.H.  
Storage Temperature:  
-20°C to 60°C (-4°F to 140°F) <80% R.H.  
cancel HOLD mode and resume measurement. In the MIN/MAX recording  
Ambient Operating Ranges:  
mode, press  
key to stop the recording. Press  
key again to resume  
recording. (Previously recorded readings are not erased).  
Backlight function mode  
Press the  
button for more than two seconds to turn on the backlight  
Press the button again for more than two seconds to turn off the backlight.  
The backlight will switch-off automatically after 30 seconds.  
GENERAL  
4. T1 T2/T1-T2 Main Display Input Selection  
The input selection determines which input is selected for the main display;  
T1 thermocouple, T2 thermocouple or the difference between the two ther-  
mocouples (T1-T2), when the thermometer is turned on it is set to T1.  
5. K/J T1 Input Type Selection (Main display)  
Display: 5 digit liquid crystal display (LCD).  
Overload: “----.-“ or “OL” is display.  
Battery: 1.5V x 4 PCS (SIZE AAA) UM-4 R03.  
Battery Life: 200 hours typical with carbon zinc battery.  
Auto power off: 30 minutes, press power key to resume operation.  
Dimensions: 160mm(H) x 83mm(W) x 38mm(D)  
Weight: Approx. 265g including batteries.  
Supplied Thermocouples (1 per input):  
1 meter (40”) type K insulated beaded wire thermocouple. Maximum insula-  
tion temperature is 482°C (900°F). Thermocouple accuracy is 1.1°C or 0.4%  
of reading (whichever is greater) from 0°C to 1250°C.  
The K/J key selects the T1 thermocouple type, when the main display is  
showing T1. When the thermometer is turned on, it is set to the type se-  
lected when the thermometer was last turned off.  
6. MIN MAX with Elapsed Time Mode (only Main display)  
Press MIN MAX key to enter the MIN MAX Recording mode, (displays the  
Maximum reading with time, Minimum reading with time and Average  
reading stored in record mode). In the this mode the automatic power-off  
feature is disabled and key, °C/°F key, REL key, SET key, Hi/Lo Limits  
key and main display T1 T2 T1-T2 key, K/J key are disabled. The beeper  
emits a tone when a new minimum or maximum value is recorded.  
SPECIFICATIONS  
ELECTRICAL  
Temperature Scale: Celsius or Fahrenheit user selectable.  
Measurement Range:  
J-TYPE -200°C to 1050°C, (-328°F to 1922°F)  
K-TYPE -200°C to 1370°C, (-328°F to 2498°F)  
Resolution: 0.1°C or 0.2°F  
MADE IN TAIWAN  
Accuracy: Accuracy is specified for operating temperatures over the range of  
18°C to 28°C (64°F to 82°F), for 1 year, not including thermocouple error.  
(0.05% rdg + 0.3°C) -50°C to 1370°C  
(0.05% rdg + 0.7°C) -50°C to -200°C  
(0.05% rdg + 0.6°F) -58°F to 2498°F  
(0.05% rdg + 1.4°F) -58°F to -328°F  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Milwaukee Portable Generator 4960 24 User Manual
MTD Lawn Mower 110 011 User Manual
MTD Lawn Mower 111 080R000 THRU 111 089R000 User Manual
MTD Tiller 214 100A User Manual
Murphy Battery Charger S10 Series User Manual
Nady Systems Stereo Amplifier UHF 16 User Manual
Newcon Optik Hunting Equipment NC 20 60X80 User Manual
NordicTrack Home Gym NTTL18903 User Manual
Onkyo Stereo Receiver HTP 340 User Manual
Oregon Scientific Clock Radio RM318PA User Manual