Operator’s Manual
Platinum™ Series
XL, 5, 10
Oxygen Concentrators Standard,
with SensO2®and HomeFill®
Invacare 5 with SensO2
Model IRC5LXO2
Invacare 5
Model IRC5LX
Invacare 10 with SensO2
Model IRC10LXO2
Invacare 10
Model IRC10LX
Dealer: This manual MUST be given to the end
user.
User: Before using this product, read this manual
and save for future reference.
For more information regarding
Invacare products, parts, and services,
Download from Www.Somanuals.com. All Manuals Search And Download.
TABLE OF CONTENTS
TABLE OF CONTENTS
ACCESSORIES ..................................................2
Injury to Persons................................................................................. 8
Unpacking............................................................................................12
Inspection............................................................................................12
Storage.................................................................................................12
PARAMETERS ............................................... 13
Introduction........................................................................................16
Set-Up..................................................................................................17
Power Switch .....................................................................................19
Flowrate...............................................................................................20
3
Download from Www.Somanuals.com. All Manuals Search And Download.
TABLE OF CONTENTS
TABLE OF CONTENTS
SECTION 6—MAINTENANCE .......................... 23
REGISTER YOUR
PRODUCT
The benefits of registering include:
1. Safeguarding your investment.
2. Ensuring long-term maintenance and servicing of your product.
3. Receiving updates with product information, maintenance tips
and industry news.
Register ONLINE at
warranty.invacare.com
Please have your model number and purchase date available to
complete your registration.
Any registration information you submit will only be used by
Invacare Corporation and protected as required by applicable laws
and regulations.
4
Download from Www.Somanuals.com. All Manuals Search And Download.
SPECIAL NOTES
SPECIAL NOTES
Signal words are used in this manual and apply to hazards or
unsafe practices which could result in personal injury or
property damage. Refer to the following table for definitions
of the signal words.
SIGNAL WORD
MEANING
Danger indicates an imminently hazardous situation
which, if not avoided, will result in death or serious
injury.
DANGER
Warning indicates a potentially hazardous situation
which, if not avoided, could result in death or
serious injury.
WARNING
Caution indicates a potentially hazardous situation
which, if not avoided, may result in property
damage or minor injury or both.
CAUTION
NOTICE
The information contained in this document is subject to
change without notice.
ꢁ DANGER
DO NOT SMOKE while using this device. Keep all
matches, lighted cigarettes or other sources of ignition
out of the room in which this product is located and away
from where oxygen is being delivered.
NO SMOKING signs should be prominently displayed.
Textiles and other materials that normally would not
burn are easily ignited and burn with great intensity in
oxygen enriched air. Failure to observe this warning can
result in severe fire, property damage and cause physical
injury or death.
5
Download from Www.Somanuals.com. All Manuals Search And Download.
SPECIAL NOTES
CAUTION
“Caution: Federal law restricts this device to sale or rental by
or on order of a physician, or any other practitioner licensed
by the law of the State in which he/she practices to use or
order the use of this device.”
Invacare recommends an alternate source of supplemental
oxygen in the event of a power outage, alarm condition or
mechanical failure. Consult your physician or equipment
provider for the type of reserve system required.
This equipment is to be used as an oxygen supplement and is
not considered life supporting or life sustaining.
6
Download from Www.Somanuals.com. All Manuals Search And Download.
LABEL LOCATION
LABEL LOCATION
IRC5LXO2 and IRC10LXO2
IRC5LX and IRC10LX
HomeFill II Compatible
HomeFill II Compatible
withSensO2
™
Operating Instructions
(See Operator's Manual)
Operating Instructions
(See Operator's Manual)
I/O
Alarms
(See Operator's Manual)
Alarms
(See Operator's Manual)
WARNING - RISK OF FIRE
WARNING - RISK OF FIRE
NO SMOKING - Keep ALL sources of ignition out of the room in
which this product is located and away from areas where oxygen
is being delivered. Textiles, oils and other combustibles are easily
ignited and burn with great intensity in oxygen enriched air.
NO SMOKING - Keep ALL sources of ignition out of the room in
which this product is located and away from areas where oxygen
is being delivered. Textiles, oils and other combustibles are easily
ignited and burn with great intensity in oxygen enriched air.
CAUTION - RISK OF ELECTRIC SHOCK
Do not remove cover. Refer servicing to qualified service
personnel.
CAUTION - RISK OF ELECTRIC SHOCK
Do not remove cover. Refer servicing to qualified service
personnel.
Flow rates below 0.5 L/min. will trigger Low Flow Alarm
(rapid audible alarm).
Flow rates below 0.5 L/min. will trigger Low Flow Alarm
(rapid audible alarm).
Set flowmeter at 3 L/min. or less when filling
HomeFill II cylinders.
Set flowmeter at 3 L/min. or less when filling
HomeFill II cylinders.
CAUTION: Federal (USA) law restricts this
CAUTION: Federal (USA) law restricts this
device to sale by or on the order of a physician.
device to sale by or on the order of a physician.
P/N 1108820
P/N 1108755
Serial Number Label is located on
the resonator intake assembly
Specification
Label is located
on the back near
the bottom
7
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 1—GENERAL GUIDELINES
SECTION 1—GENERAL GUIDELINES
In order to ensure the safe installation, assembly and
operation of the Platinum concentrator these instructions
MUST be followed.
ꢀ WARNING
This section contains important information for the safe
operation and use of this product.
To Reduce the Risk of Burns,
Electrocution, Fire or Injury to Persons
ꢁ DANGER
Risk of electric shock. DO NOT disassemble. Refer servicing
to qualified service personnel. No user serviceable parts.
Avoid using while bathing. If continuous usage is required
by the physician’s prescription, the concentrator MUST be
located in another room at least 7 ft from the bath.
DO NOT come in contact with the concentrator while wet.
DO NOT place or store product where it can drop into
water or other liquid.
DO NOT reach for product that has fallen into water.
Unplug IMMEDIATELY.
If the concentrator has a damaged cord or plug, if it is not
working properly, if it has been dropped or damaged, or
dropped into water, call qualified technician for
examination and repair.
A spontaneous and violent ignition may occur if oil, grease
or greasy substances come in contact with oxygen under
pressure. These substances MUST be kept away from the
oxygen concentrator, tubing and connections, and all
other oxygen equipment. DO NOT use any lubricants
unless recommended by Invacare.
Avoid creation of any spark near medical oxygen
equipment. This includes sparks from static electricity
created by any type of friction.
8
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 1—GENERAL GUIDELINES
Operating Information
For optimum performance, Invacare recommends that each
concentrator be on and running for a minimum of 30 minutes
at a time. Shorter periods of operation may reduce maximum
product life.
Keep the oxygen tubing, cord, and unit out from under such
items as blankets, bed coverings, chair cushions, clothing and
away from heated or hot surfaces, including space heaters,
stoves and similar electrical appliances.
DO NOT move or relocate concentrator by pulling on the
power cord.
NEVER drop or insert any object or liquid into any opening.
Invacare recommends that Crush‐Proof oxygen tubing be
used with this product and not exceed 15.2 m (50 ft) in length.
There are no user serviceable parts. This does not include
normal maintenance items. See maintenance section for user
maintenance items.
A product should NEVER be left unattended when plugged
in. Make sure the Platinum is Off when not in use.
Close supervision is necessary when this product is used near
children or physically‐challenged individuals.
Additional monitoring or attention may be required for
patients using this device who are unable to hear or see
alarms or communicate discomfort.
DO NOT connect the concentrator in parallel or series with
other oxygen concentrators or oxygen therapy devices.
Radio Frequency Interference
This equipment has been tested and found to comply with
EMC limits specified by IEC/EN 60601‐1‐2. These limits are
designed to provide a reasonable protection against
electromagnetic interference in a typical medical installation.
9
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 1—GENERAL GUIDELINES
Other devices may experience interference from even the low
levels of electromagnetic emissions permitted by the above
standards. To determine if the emissions from the Platinum is
causing the interference, turn the Platinum Off. If the
interference with the other device(s) stops, then the Platinum
is causing the interference. In such rare cases, interference
may be reduced or corrected by one of the following
measures:
• Reposition, relocate, or increase the separation
between the equipment.
Connect the equipment into an outlet on a circuit different
from that to which the other device(s) is connected.
10
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 2—FEATURES
SECTION 2—FEATURES
Oxygen Purity Indicator Lights
(SensO2 model only)/ Fault
and Power Indicator Lights
(LX model only)
Oxygen Outlet
Flowmeter
Power Switch
Circuit Breaker
Elapsed Time Meter
Cabinet Filter
(one of two shown)
ACCESSORIES (NOT SHOWN):
Remote Humidifier/Flowmeter Accessory -
IRCRF5 (Platinum 5 and XL Only)
PreciseRX™ Pediatric Humidifier/Flowmeter
Accessory - IRCPF16 (Platinum 5 and XL Only)
REAR VIEW
HomeFill home oxygen compressor - IOH200
NOTE: This outlet fitting is to be used only
for filling oxygen cylinders with the
HomeFill home oxygen compressor. The
outlet fitting does not affect concentrator
performance. Refer to the HomeFill ownerʹs
manual, part number 1100873, for
connection and operating instructions.
When not in use, the plug provided with the
concentrator should be inserted into the
outlet fitting. For more information about the
HomeFill, contact your Invacare dealer.
Outlet
Fitting
Power
Cord
11
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 3—HANDLING
SECTION 3—HANDLING
The concentrator should ALWAYS be kept in the upright
position to prevent cabinet damage while being transported.
If the concentrator is to be reshipped by common carrier,
additional cartons are available from Invacare.
Unpacking
NOTE: For this procedure, refer to FIGURE 3.1.
1. Check for any obvious
damage to the carton or
its contents. If damage is
evident, notify the
carrier, or your local
dealer.
2. Remove all loose packing
from the carton.
3. Carefully remove all the
components from the
carton.
NOTE: Unless the oxygen concentrator is to be used immediately,
retain containers and packing materials for storage until use of the
concentrator is required.
Inspection
Inspect/examine exterior of the oxygen concentrator for nicks,
dents, scratches or other damages. Inspect all components.
Storage
1. Store the repackaged oxygen concentrator in a dry area.
2. DO NOT place objects on top of repackaged concentrator.
12
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 4—TYPICAL PRODUCT PARAMETERS
SECTION 4—TYPICAL PRODUCT
PARAMETERS
Electrical Requirements:
120 VAC ± 10% (132 VAC/108 VAC), 60
Hz
Rated Current Input:
Platinum 5 and XL: 4.3 A
Platinum 10: 4.7 A
Sound Level:
Altitude:
Platinum 5 and XL: 50 dB Average
Platinum 10: 58 dB Average
Platinum 5 and XL - Up to 6,000 ft (1828
meters) above sea level without degradation
of concentration levels. From 6,000 ft (1828
meters) to 13,129 ft (4000 meters) below
90% efficiency
Platinum 10 - Up to 4,000 ft (1230 meters)
above sea level without degradation of con-
centration levels. From 4,000 ft (1230
meters) to 13,129 ft (4000 meters) below
90% efficiency
*Oxygen Output
Concentration Levels:
All 5LXO2/5LX models 87% to 95.6% for
flows ranging 0.5 to 5 L/min and all
10LXO2/10LX models 87% to 95.6% for
flows ranging 2 L to 10 L/min
Maximum Outlet Pressure:
Platinum 5 and XL: 5 psi ± 0.5 psi (34.5 kPa
± 3.45 kPa)
Platinum 10: 9 psi ± 0.5 psi (62.0 kPa ± 3.45
kPa)
*NOTE: Stated concentration levels achieved after initial warm‐up
period (approximately 30 minutes).
13
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 4—TYPICAL PRODUCT PARAMETERS
Flow Range:
Platinum 5 and XL - 0.5 to 5 L/min For
flowrates less than 1 L/min, we recommend
the use of the Invacare Pediatric Flowmeter
IRCPF16 Accessory.
Platinum 10 - 2 L/min to 10 L/min
Low Flow Alarm:
Platinum 5 - 0 L/min to 0.5 L/min,
Platinum 10 - 0 L/min to 1 L/min
Rapid Audible Alarm Beeping (No
Accessories Connected)
Average Power
Consumption:
Platinum 5 and XL: 400W
Platinum 10: 585W
Pressure Relief Mechanism
Operational at:
35 psi ± 3.5 psi (241 kPa ± 24.1 kPa)
Change in maximum recom-
mended flow when back pres-
sure of 7kPa is applied:
0.7 L/min
Filters:
Cabinet (2), Outlet HEPA and Compressor
Inlet.
Safety System:
Current overload or line surge shutdown.
High temperature compressor shutdown.
High Pressure Alarm w/compressor shut-
down. Low Pressure Alarm w/compressor
shutdown. Battery Free Power Loss Alarm.
SensO2 Oxygen System (SensO2 Model)
Low Flow Alarm.
Width:
Height:
Depth:
Weight:
18 3/8 in ± 3/8 in (46.7 cm ± 1 cm)
26 3/8 in ± 3/8 in (67.0 cm ± 1 cm)
14 3/8 in ± 3/8 in (36.5 cm ± 1 cm)
Platinum 5 and XL: 52 lbs ± 2 lbs (23.6 kg ±
1 kg)
Platinum 10: 53 lbs ± 2 lbs (24.0 kg ± 1 kg)
Shipping Weight:
Platinum 5 and XL: 58 lbs ± 2 lbs (26.3 kg ±
1 kg)
Platinum 10: 60 lbs ± 2 lbs (27.2 kg ± 1 kg)
14
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 4—TYPICAL PRODUCT PARAMETERS
Operating Ambient
Temperature:
50°F - 95°F (10°C - 35°C) at 20-60% relative
humidity
Exhaust:
Platinum 5 and XL: Less than Ambient
+ 35°F (+19°C).
Platinum 10: Less than Ambient + 45°F
(+25°C)
Oxygen Output Temperature: Platinum 5 and XL: Less than Ambient + 4°F
(+2°C)
Platinum 10: Less than Ambient + 8°F
(+4.4°C)
Cabinet:
Impact Resistant flame-retardant plastic cab-
inet that conforms to UL 94-V0
Regulatory Listing:
Double Insulated Product ETL certified
complying with UL 1097 ETL certified to
CSA C22.2 No. 68
Electrical:
No extension cords.
Placement:
No closer than three inches from any wall,
furniture, draperies, or similar surfaces.
Tubing:
7 ft cannula with a maximum 50 ft of
Crush-Proof Tubing (DO NOT pinch).
Platinum 10: Recommended use up to 50 ft
high flow tubing with high flow cannula at all
flow rates.
Relative Humidity:
Time of Operation:
20 to 60%
Up to 24 hours per day.
Recommended Storage and
Shipping Temperature:
-20°F to 150°F (-29°C to 65°C), at 15-95%
relative humidity
15
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 5—OPERATING INSTRUCTIONS
SECTION 5—OPERATING
INSTRUCTIONS
Introduction
Your oxygen concentrator is intended for individual use
indoors. It is an electronically operated device that separates
oxygen from room air. It provides high concentration of oxygen
directly to you through a nasal cannula. Clinical studies have
documented that oxygen concentrators are therapeutically
equivalent to other types of oxygen delivery systems.
Your provider will show you how to use your oxygen
concentrator. He/She should be contacted with any questions
or problems regarding your oxygen concentrator. This
owner’s manual will tell you about your concentrator and will
serve as a reference as you use your concentrator.
Select a Location
ꢀ WARNING
NEVER block the air openings of the product or place it on a
soft surface, such as a bed or couch, where the air opening may
be blocked. Keep the openings free from lint, hair and the like.
Keep unit at least 3 inches away from walls, draperies,
furniture, and the like.
You may select a room where using your oxygen concentrator
would be most convenient. Your concentrator can be easily
rolled from room to room on its casters.
The air intake of the unit should be located in a well ventilated
area to avoid airborne pollutants and/or fumes. DO NOT
place in a closet.
16
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 5—OPERATING INSTRUCTIONS
Set-Up
Plug in Power Cord
Plug in power cord to an electrical outlet.
ꢀ POLARIZED PLUG WARNING
As a safety feature, this appliance may have a polarized plug
(one blade is wider than the other). This plug will fit in a
polarized outlet only one way. If the plug does not fit fully in the
outlet, reverse the plug. If it still does not fit, contact a qualified
electrician. DO NOT attempt to defeat this safety feature.
Connect Humidifier (If So Prescribed)
ꢀ WARNING
DO NOT overfill humidifier.
DO NOT reverse the oxygen input and output connections.
Water from the humidifier bottle will travel through the
cannula back to the patient.
NOTE: For Platinum 5, XL ‐ recommended humidifiers are:
Invacare 3260, 003‐40, or 006‐40. For Platinum 10, recommended
humidifiers are high flow: 7900‐10, 7900‐25.
1. Remove cap and fill humidifier bottle with distilled water
to the level indicated by the manufacturer.
2. Replace the humidifier cap and securely tighten.
Humidifier Bottle
FIGURE 5.1 Filling the Humidifier
17
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 5—OPERATING INSTRUCTIONS
Humidifier Bottle
FIGURE 5.2 Humidifier Compartment
3. Remove the filter access door located on the side of the
concentrator. Insert a flathead screwdriver in the plate
groove on the top edge of the filter access door and gently
4. The humidifier adapter is next to the inlet filter. Pull up
and remove the humidifier bottle adapter. Refer to Detail
5. Attach it to the humidifier by turning the wing nut on the
humidifier bottle counterclockwise until it is securely
6. Place the humidifier bottle in the humidifier compartment.
Plate Groove (Insert
Flat Head Screwdriver
DETAIL
Here)
“B”
Filter
Access
Humidifier
Door
Bottle
Adapter
DETAIL
“A”
DETAIL “C”
Filter
Access
Door
Humidifier Bottle
Adapter
FIGURE 5.3 Attaching the Humidifier Bottle Adapter
18
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 5—OPERATING INSTRUCTIONS
7. Attach oxygen tubing from the humidifier bottle to
oxygen outlet connector on the oxygen concentrator
8. Attach the cannula/patient supply tubing to the
humidifier bottle outlet (FIGURE 5.4).
9. After assembly, ensure that oxygen is flowing through the
cannula.
Oxygen Outlet
Connector
Oxygen
Tubing
Humidifier Outlet
FIGURE 5.4 Attaching the Oxygen Tubing
Power Switch
NOTE: For this procedure, refer to FIGURE 5.5 on page 19.
1. Press power switch to on position. All the panel lights and
the audible alarm will come on for one second, indicating
that the unit is functioning properly.
On/Off (I/O)
Power Switch
19
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 5—OPERATING INSTRUCTIONS
Flowrate
NOTE: For this procedure, refer to FIGURE 5.6 on page 20.
1. Turn the flowrate knob to the setting prescribed by your
physician or therapist.
CAUTION
Never set the flow greater than 5 L/min on the IRC5.
Never set the flow greater than 10L/min on the IRC10.
ꢀ WARNING
DO NOT change the L/min setting on the flowmeter unless a
change has been prescribed by your physician or therapist.
NOTE: To properly read the flowmeter, locate the prescribed flowrate
line on the flowmeter. Next, turn the flow knob until the ball rises to
the line. Now, center the ball on the L/min. line prescribed.
2. If the flowrate on the flowmeter ever falls below 0.5 L/min
for Platinum 5 (1 L/min for Platinum 10) for more than
about one minute, the LOW FLOW alarm will be
triggered. This is a rapid beeping of the audible alarm.
Check your tubing or accessories for blocked or kinked
tubing or a defective humidifier bottle. After rated flow is
restored, more than 0.5 L/min for Platinum 5 (1 L/min for
Platinum 10), the LOW FLOW audible alarm will go off.
NOTE: The use of some accessories such as the pediatric flowstand
and the HomeFill compressor will deactivate the Low Flow Alarm.
Flow Knob
Ball
20
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 5—OPERATING INSTRUCTIONS
Front Panel Indicators - Units with SensO2
NOTE: For this procedure, refer to FIGURE 5.7.
The SensO2 feature monitors the purity of oxygen generated by
the oxygen concentrator. If purity falls below factory preset
standards, indicator lights on the control panel will illuminate.
OXYGEN PURITY
Auto Shut - Down
RED
Use Backup
Call Supplier
Below Normal
YELLOW
GREEN
Normal
O2
Initial Startup of the Concentrator
NOTE: Concentrator may be used during the initial start warm‐up
time (approximately 30 min.) while waiting for the O2 purity to
reach maximum.
LABEL
SYMBOL
O2 PURITY
INDICATOR LIGHTS
(LED)
SYSTEM OKAY
O2 over 85% (± 2%)
GREEN Indicator Light
O2
YELLOW Indicator light
A. YELLOW Solid
B. YELLOW Flashing Sensor Failure
Call a qualified technician. Have a
back-up supply of oxygen ready.
O2 Between 73% (± 3%)
to 85% (±2%)
SYSTEM FAILURE
RED Indicator Light
O2 Below 73% (±3%)
Continuous Audible Alarm
Sieve-GARD Compressor Shut-
down. Call a qualified technician.
Switch to a back-up supply of oxy-
gen immediately.
21
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 5—OPERATING INSTRUCTIONS
Front Panel Indicators - Units without
SensO2
NOTE: For this procedure, refer to FIGURE 5.8.
Auto Shut - Down
RED
Use Backup
Call Supplier
I/O
Normal
GREEN
Initial Startup of the Concentrator
NOTE: Concentrator may be used during the initial start warm‐up
time (approximately 30 min). O2 purity will build to a maximum
during this period.
LABEL
SYMBOL
STATUS
INDICATOR LIGHTS
(LED)
SYSTEM OKAY
GREEN Indicator Light
I/O
SYSTEM FAILURE
RED Indicator light
Continuous Audible Alarm
Sieve-GARD™Compressor
Shutdown
Call a qualified technician. Switch
to a back-up supply of oxygen
immediately.
22
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 6—MAINTENANCE
SECTION 6—MAINTENANCE
ꢀ WARNING
The Invacare concentrators are specifically designed to
minimize routine preventive maintenance. Only qualified
personnel should perform preventive maintenance on the
concentrator.
Power should be disconnected before beginning preventive
maintenance on the Invacare concentrator.
NOTE: At a minimum, preventive maintenance MUST be
performed according to the maintenance record guidelines. In places
with high dust or soot levels, maintenance may need to be performed
Routine Maintenance
NOTE: The following routine maintenance should be performed
more frequently. See recommended intervals in each section and
preventive maintenance record.
Cleaning the Cabinet Filter
CAUTION
DO NOT operate the concentrator without the filter installed.
NOTE: For this procedure, refer to FIGURE 6.1 on page 24.
NOTE: There are two cabinet filters one located on each side of the
cabinet.
1. Remove each filter and clean at least once a week
depending on environmental conditions.
NOTE: Environmental conditions that may require more frequent
cleaning of the filters include but are not limited to: high dust,
smoking, air pollutants, etc.
23
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 6—MAINTENANCE
2. Clean the cabinet filters with a vacuum cleaner or wash in
warm soapy water and rinse thoroughly.
3. Dry the filters thoroughly before reinstallation.
Cabinet Filter
(located on both
sides of
concentrator)
Cleaning the Cabinet
ꢀ WARNING
UNPLUG the concentrator when cleaning. To avoid electrical
shock, DO NOT remove cabinet.
Clean the cabinet with a mild household cleaner and
non‐abrasive cloth or sponge.
Cleaning the Humidifier
NOTE: To clean the oxygen humidifier, follow the instructions
provided by the manufacturer. If none are provided, follow these
STEPS:
1. Clean the humidifier every day.
2. Wash it in soapy water and rinse it with a solution of ten
parts water and one part vinegar.
3. Rinse thoroughly with hot water and refill with distilled
water to the level shown on the humidifier.
24
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 7—TROUBLESHOOTING GUIDE
SECTION 7—TROUBLESHOOTING
GUIDE
SYMPTOM:
PROBABLE CAUSE:
SOLUTION:
Alarm:
SHORT BEEPS,
LONG PAUSE
Main Power Loss:
1. Power cord NOT
plugged in.
1. Insert plug into outlet.
2. No power at outlet.
2. Inspect house circuit break-
ers or fuses. If problem recurs,
use a different outlet.
Concentrator
NOT operating,
power switch ON.
3. Tripped circuit breaker. 3. Call Service Provider.
Beep......Beep......
Alarm:
System Failure:
1a. Remove and clean cabinet
CONTINUOUS
1. Unit overheating due to filters.
blocked air intake.
1b. Move oxygen concentrator
at least three inches away from
walls, draperies or furniture.
Concentrator
NOT operating,
Power Switch ON.
2. Insufficient power at
outlet.
2. DO NOT use extension
cords. Move to another elec-
trical outlet or circuit.
Beeeeeeeep....
3. Internal repairs
required
3. Call Service Provider.
Alarm:
Low Flow Alarm:
1. Inspect for kinks or
RAPID
1. Kinked or blocked tub- blockages. Correct, clean or
ing, cannula or humidifier. replace item. Once
corrected, turn power OFF
Beep....Beep...
Beep...Beep
for 60 seconds and then turn
power back ON.
2. Flowmeter set too low: 2. Reset flowmeter to pre-
0.5 L/min (or less) for
scribed flowrate. If prescribed
Platinum 5; 1.0 L/min (or flowrate is below 0.5 L/min,
less) for Platinum 10.
use the Pediatric Flowmeter
IRCPF16 (Only with Platinum
5 and XL).
26
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 7—TROUBLESHOOTING GUIDE
SYMPTOM:
PROBABLE CAUSE:
SOLUTION:
YELLOW or RED 1. Low oxygen purity.*
1. Clean or Replace filters.
LIGHT Illuminated.
2. Kinked or blocked tub- 2. Inspect for kinks or block-
ing, cannula or humidifier.* ages. Correct, clean or replace
item. Once corrected, turn
Alarm:
CONTINUOUS
On red light only.
power OFF for 60 seconds and
then turn power back ON.
3. Flowmeter set too low: 3. Check flowmeter is set to
* Only applies to
5LXO2 and
10LXO2 SensO2
models.
0.5 L/min (or less) for
Platinum 5; 1.0 L/min (or
less) for Platinum 10.
1.0 L/min. or more.
4. Unit overheating due to 4.a. Remove and clean cabinet
blocked air intake.
filters.
4b. Move concentrator at least
three inches from walls, drap-
eries, and furniture.
5. Insufficient power at
outlet.
5. DO NOT use extension
cords. Move to another elec-
trical outlet or circuit.
6. Internal repairs
required.
6. Call Service Provider.
GREEN LIGHT
with YELLOW
Light Flashing.
1. Internal repairs
required.
1. Call Service Provider.
27
Download from Www.Somanuals.com. All Manuals Search And Download.
NOTES
NOTES
29
Download from Www.Somanuals.com. All Manuals Search And Download.
LIMITED WARRANTY
LIMITED WARRANTY
NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY
WITH FEDERAL LAW APPLICABLE TO PRODUCTS
MANUFACTURED AFTER JULY 4, 1975.
This warranty is extended only to the original purchaser who purchases
this product when new and unused from Invacare Corporation or a
dealer. This warranty is not extended to any other person or entity and it
is not transferable or assignable to any subsequent purchaser or owner.
Coverage under this warranty will end upon any such subsequent sale or
other transfer of title to any other person. This warranty gives you specific
legal rights and you may also have other legal rights which may vary from
state to state.
Invacare Corporation warrants its Platinum XL Oxygen Concentrators
when purchased new and unused to be free from defects in materials and
workmanship for a period of five years from date of purchase from
Invacare or a dealer, with a copy of the seller’s invoice required for
coverage under this warranty.
Invacare Corporation warrants its Platinum 10 Oxygen Concentrators
when purchased new and unused to be free from defects in materials and
workmanship for a period of three years from date of purchase from
Invacare or a dealer, with a copy of the seller’s invoice required for
coverage under this warranty.
DO NOT OPEN OR ATTEMPT TO SERVICE THE 4-WAY VALVE
BODY. This will void any and all warranty applicable to the valve body. If
within such warranty periods any such product shall be proven to Invacare
Corporation's satisfaction to be defective, such product shall be repaired
or replaced, at Invacare Corporation's option. This warranty only applies
to the labor for repairs performed by the Invacare Service Department or
Invacare Authorized Service Centers. It does not apply to the labor
performed by the purchaser or user. This warranty does not include
normal wear and tear or shipping charges. Invacare Corporation's sole
obligation and your exclusive remedy under this warranty shall be limited
to such repair or replacement. Routine maintenance items, such as filters,
are excluded from this warranty.
30
Download from Www.Somanuals.com. All Manuals Search And Download.
LIMITED WARRANTY
For warranty service, please contact Invacare Corporation's Service
Department at the toll free number on the back cover. Upon receiving
notice of an alleged defect in a product, Invacare Corporation will issue a
serialized return authorization. It shall then be the responsibility of the
purchaser to return the entire unit or remove, at purchaser's cost, the
defective component part(s) identified, pack the component part(s) in a
manner to avoid shipping damage and to ship the component part(s) to
either Invacare Corporation's plant or service center as specified by
Invacare Corporation in advance. Defective component part(s) MUST be
returned for warranty inspection using the serial number as identification
within thirty (30) days of return authorization date. DO NOT return
products to our factory without prior consent. C.O.D. shipments will be
refused; please prepay shipping charges.
LIMITATIONS AND EXCLUSIONS: THE FOREGOING WARRANTY
SHALL NOT APPLY TO PRODUCTS SUBJECTED TO NEGLIGENCE,
ACCIDENT, IMPROPER OPERATION, MAINTENANCE OR STORAGE,
SOOT OR SMOKE-FILLED ENVIRONMENTS, OR OTHER THAN
NORMAL APPLICATION, USE OR SERVICE, OR TO PRODUCTS
MODIFIED WITHOUT INVACARE CORPORATION'S EXPRESS
WRITTEN CONSENT (INCLUDING, BUT NOT LIMITED TO,
MODIFICATION THROUGH THE USE OF UNAUTHORIZED PARTS
OR ATTACHMENTS) OR TO PRODUCTS DAMAGED BY REASON OF
REPAIRS MADE TO ANY COMPONENT WITHOUT THE SPECIFIC
CONSENT OF INVACARE CORPORATION OR TO PRODUCTS
DAMAGED BY CIRCUMSTANCES BEYOND INVACARE
CORPORATION'S CONTROL.
THE FOREGOING EXPRESS WARRANTY IS EXCLUSIVE AND IN
LIEU OF ANY OTHER WARRANTIES WHATSOEVER, WHETHER
EXPRESS OR IMPLIED, INCLUDING THE IMPLIED WARRANTIES OF
MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE,
AND THE SOLE REMEDY FOR VIOLATIONS OF ANY WARRANTY
WHATSOEVER, SHALL BE LIMITED TO REPAIR OR REPLACEMENT
OF THE DEFECTIVE PRODUCT PURSUANT TO THE TERMS
CONTAINED HEREIN. THE APPLICATION OF ANY IMPLIED
WARRANTY WHATSOEVER SHALL NOT EXTEND BEYOND THE
DURATION OF THE EXPRESS WARRANTY PROVIDED HEREIN.
INVACARE SHALL NOT BE LIABLE FOR ANY CONSEQUENTIAL OR
INCIDENTAL DAMAGES WHATSOEVER.
SOME STATES DO NOT ALLOW THE EXCLUSION OR LIMITATION
OF INCIDENTAL OR CONSEQUENTIAL DAMAGE, OR LIMITATION
OF HOW LONG AN IMPLIED WARRANTY LASTS, SO THE ABOVE
EXCLUSION AND LIMITATION MAY NOT APPLY TO YOU.
THIS WARRANTY SHALL BE EXTENDED TO COMPLY WITH
STATE/PROVINCIAL LAWS AND REQUIREMENTS.
31
Download from Www.Somanuals.com. All Manuals Search And Download.
This product may be manufactured under one or more of the
following patents: US 5,906,672 and US 5,917,135. Other
patents pending.
Invacare Corporation
USA
Trademarks are identified by ™ and ®. All
trademarks are owned by or licensed to
Invacare Corporation or its subsidiaries
unless otherwise noted.
One Invacare Way
Elyria, Ohio USA
44036-2125
800-333-6900
Technical Services
800-832-4707
© 2009 Invacare Corporation
Part No 1118353
Rev J - 4/09