Pharmaceutical Refrigerator  
Operation Manual  
HYC-610  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Safety Precautions  
The Haier Pharmaceutical Refrigerator unit is designed for storage of  
whole blood and related products at constant temperature in clinical,  
pharmaceutical, research, and other laboratory applications.Please  
carefully read this manual, and follow the instructions to properly operate  
the unit.  
Example of labels:  
This manual covers all aspects of  
This symbol suggests precaution be  
installing and operating the unit. It also  
provides information that can prevent a  
user from being injured. “Caution” and  
“Warning” suggest two levels of safety  
implications. Please follow them carefully.  
followed.  
This symbol means that an action is  
strictly prohibited.  
This symbol designates procedures  
to be followed.  
Please keep this manual for convenient  
reference  
Warning  
This action may cause fatal injury or severe  
damage.  
Environmental Conditions  
This unit is designed for applications in the following environmental conditions. It is safe to  
operate the unit at these or more favorable conditions:  
1)Indoor usage  
2)Maximum relative humidity of 80 percent at an ambient of less than 31 . Maximum allowable  
relative humidity will decrease as the ambient temperature rises.  
3)Voltage variation within  
10 percent of nominal voltage supply  
4)Factory approved abnormal voltage range  
5)Instantaneous over-voltage complies with second grade for installation of equipment. For the  
main voltage supply, the minimum and normal supply is second level.  
6)Environment complying with IEC664 standard, Grade 2 Contamination.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Safety Precautions  
After-Sale Support and Warning Instruction  
Warning  
Dear Customer,  
As any products might be subject to a possibility of failure, we encourage you to routinely observe the  
operation of the refrigerator and any change of the stored products. Should there be any changes,  
please attempt to follow the instruction in the manual for resolution. If the problem persists, please notify  
your local Haier service center for quick service. This is to reduce the chance of product loss.  
Do not use this refrigerator in outdoor  
conditions. If the unit is rained on, there is a  
possibility of electric shock or electrocution.  
During loading and unloading products from  
the unit, do not pull out two drawers at the  
same time. If there are heavy loads in the  
drawers, the unit may tilt over and fall  
causing injury.  
The installation of this refrigerator unit should  
only be performed by a certified technician.  
Improper installation may lead to refrigerant  
gas leak or liquid leak, electric shook or fire.  
After Sale Service  
Do not directly apply water into the  
refrigerator as the moisture may cause  
electrical shortage and electrical shock.  
1.From the date this unit is sold, the entire unit is warranted for one year. The key components such as  
the evaporator, condenser, filter dryer, compressor, capillary tube, fans, and blowers are guaranteed for  
three years.  
The unit should be situated on a solid surface  
or ground. If the installation is not on a solid  
foundation, the unit can tilt or tip over to fall  
causing injury or property damage.  
Do not allow unqualified personnel to  
dismantle, repair or modify the refrigerator.  
Such action may cause fire or injury of  
personnel.  
2.Warranty certificate: When you need to have service done or to inquire about the unit, please provide  
this warranty certificate and proof of purchase to your local Haier service center and please keep this  
document in a safe place.  
Do not place the unit in a place where it is  
damp or is subject to water spray. Damage  
to the insulation of the unit may cause  
electrical shock or electrocution.  
Should there be malfunctions of the unit,  
unplug the power cord from the power  
supply. Continuous operation of the unit  
abnormally may cause fire or electrical  
shock.  
The unit must be connected to a dedicated  
power supply line. Using an extension cord  
or a power strip may cause fire.  
When used for storage of poisonous  
products, or products that are radiant and  
products that can be harmful to public  
health, the unit needs to be placed in a safe  
location. Improperly using the unit with  
these types of products may cause harm to  
the user's health or to the environment.  
The power plug must be plugged securely  
into a clean power outlet. A dusty power  
outlet may cause fire.  
The power receptacle must be equipped with  
a
ground prong.  
receptacle does not include  
If an intended power  
ground,  
a
a
Decommission of old refrigerators must be  
performed by professional workers. Do not  
discard unused refrigerators in places  
where children could play with them and  
accidentally get locked inside.  
certified electrician must install a ground wire  
before the refrigerator can be plugged to the  
power outlet.  
Never use gas line, water pipe, telephone line,  
or lightening rod for grounding wire for the  
unit. Such installation may cause  
electrocution.  
If there is a need to extend the power cord  
for the unit, the cross area of the extended  
cord conductor must be no less than 1.5  
square millimeter.  
When moving the  
Do not probe into the air ventilation holes.  
probing object may accidentally touch moving  
parts and cause injury.  
A
equipment, be careful not to run over the  
power cord. If the power cord is damaged,  
it should be repaired with specialty  
conductors or with a repair kit purchased  
from Haier.  
Do not place volatile or flammable materials  
into the refrigerator.  
Storage of such  
materials in the refrigerator may cause an  
explosion or fire.  
Should there be any abnormality about the  
temperature, or the temperature alarm  
sounds, transfer the stored products inside  
the refrigerator to a safe place immediately.  
This will reduce the chance of loss of  
properties.  
Before any repair work or maintenance or  
inspection is done, the power cord must be  
unplugged from the power supply. This is to  
prevent any risk of potential injury or electrical  
shock.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Descriptions of Refrigerator Parts and  
Their Functions  
Specification  
Description  
Pharmaceutical Refrigerator For Medical Application  
HYC-610  
Model Number  
External Dimension  
W780*D840*H1960mm  
Control Panel  
Light Power Switch  
Internal Dimension  
W680*D620*H1400mm  
610L  
Effective Storage  
Volume  
External Wall Surface  
Powder Coated Cold Rolled Steel  
Stainless Steel  
Internal Wall Surface  
Door  
Insulation  
Electric heater heated glass door  
Hard Polyurethane foam (No Fluoride)  
Drawer/Shelves  
6 Plastic shelves  
Refrigeration system  
Forced air cooling  
Hermetically sealed  
Compressor  
Condenser  
Evaporator  
Refrigerant  
Defrost system  
Finned coil condenser  
Finned coil evaporator  
R-134a  
Automatically Forced defrost system  
Electronic Controller  
Temperature controller  
High Temperature Alarm  
Low Temperature Alarm  
Door Ajar Alarm  
Flashing Alarm Indicator, alarm buzzer, remote alarm contact  
Flashing Alarm Indicator, buzzer alarm without time delay, remote alarm contact  
Alarm for door opening after 10 minutes delay  
Non-Volatile memory storage  
Memory Device  
Florescent Light  
15 W X 1  
Door Lock  
Temperature Recorder Including one box of recorder chart paper, and 4 pieces of 9 V battery (optional)  
Door Alarm Switch  
One set of key, one certificate of warranty, one copy of operation manual,  
one plastic bag  
Accessory items  
Automatic Temperature (Optional)  
Recorder  
Weight  
204 Kg  
Rechargeable battery  
12V/7.0AH  
Operating Parameters  
Caster  
Model  
HYC-610  
Sensor Bottle Temperature  
Design Ambient Temperature  
Rated Power  
2-8  
Due to continuous improvements on the products, your actual Haier refrigerator may not resemble  
the depicted image thus published in the catalog. We apologize for that.  
+5 ~+35  
490W  
72 Hours With Fully chargeable battery  
Sustainable Alarm Time  
Note: the design parameters of this refrigerator may change without notice.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Descriptions of Refrigerator Parts and  
Their Functions  
Maintenance  
1)Control Panel  
Warning  
The control panel on the Haier blood bank shows the actual storage temperature inside the refrigerator.  
You can use the push buttons on the panel to adjust the temperature set point and test the alarm  
functions.  
Before any inspection or maintenance work is performed, the refrigerator's power plug should be  
disconnected from the power supply socket. This is to prevent any potential electrical shock or injury.  
During the maintenance work, do not breathe the dust and aerosols near the unit; they might be  
harmful to your health.  
2)Light Power Switch  
You can use the switch to turn on and off the interior lights of the refrigerator.  
Cleaning the refrigerator  
3)Condensation Heater Control  
1.The refrigerator should be cleaned once a month. Regular cleaning can keep the unit looking new.  
2.Use a dry, soft cloth to clean the dust off the interior and exterior. If necessary, use a soft cloth with a  
solution of water and mild detergent to wipe the unit.  
You can use this switch (located inside the back of the refrigerator behind a cover) to turn on the  
heater that reduces the condensation build up. “1” means power on and “0” means power off.  
3.After the cleaning, use a dry cloth to wipe off any solution residue off the surfaces.  
4.Do not pour water directly into the refrigerator. By doing so, the water can damage the insulation  
materials and cause problems.  
4)Battery Power Switch  
The switch controls the rechargeable battery that powers the alarm system in case of the loss of main  
power. “1” is on and “0” is off.  
5.Parts in the refrigeration system for this refrigerator are completely sealed. They do not require any  
lubricants.  
5)Door Lock  
6.The air filter for the condenser should be cleaned at least once within a 90 day period. The procedure  
is as follows.  
You can use this feature to lock the refrigerator.  
a.Remove the front louver panel for the refrigerator.  
6)Door Handle  
b.Remove the air filter located in front of the condenser; wash it in water.  
c.After cleaning the filter, install it in front of the condenser and the louver panel.  
For ease of opening and closing the door.  
7)Casters  
Changing the light bulb starter  
The casters underneath the unit provide the convenience of moving the unit around.  
The florescent light bulb is located in the upper front area of the refrigerator. When changing the light  
bulb, please perform the following steps.  
8)Automatic Temperature Recorder (optional)  
1.Unplug the refrigerator from the power supply.  
The refrigerator is equipped with an automatic temperature recorder. Its paper chart runs for a full turn  
in 7 days, so the storage temperature can be permanently recorded. The chart paper must be  
changed once every 7 days.  
2.Remove the left panel and the cover for the electrical control box.  
3.Remove the defective starter.  
4.Install a new starter. Make sure it is securely in position.  
5.Install the cover for the electrical box and the left panel.  
6.Plug the power plug back into the system.  
9)Leveling Leg  
The unit must be level to operate. You can adjust the screws on the leveling leg underneath the  
refrigerator to adjust the height.  
Purchasing chart paper for temperature chart recorder  
The chart recorder papers are imported chart papers designed for the recorder. The papers that are  
supplied with the unit usually last about half of a year. When you are running short of the chart paper,  
please contact Haier to purchase more. The lead time to deliver the charts is 15 days after receiving the  
payment. Please call Haier for details.  
10)Shelves  
The drawers and shelves are designed to conveniently store and unload the blood samples.  
11)Interior Lights  
15 Watts light hulb.  
Purchasing backup battery for temperature recorder  
You can purchase a high quality battery from a store for the same size and voltage (9V). You may also  
call Haier to purchase a high quality replacement battery. The lead time to deliver the battery is 15 days  
after receiving the payment. Please call Haier for details.  
12)Door Ajar Alarm  
When the door is open for 10 minutes, the door ajar alarm will sound. You should reduce the open  
door time to stabilize the refrigerator temperature.  
Changing fuse breaker  
If there is a burnt out fuse on a PC board, please contact Haier's after sales support department.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Location of Refrigerator  
Proper usage of the Refrigerator  
Your refrigerator should be placed at a location that satisfies the following conditions. By doing so, the  
optimum operating results can be achieved.  
During the initial startup and continuous usage of the refrigerator, the following procedures should be  
followed.  
1.Power the refrigerator in a dedicated power socket. The voltage requirement is 220 VAC and 50  
hz. When started, do not place any products in the refrigerator.  
1)Firm, level floor surface  
Installing the unit on a firm and flat floor surface  
reduces the chance of excessive noise and  
2.When the unit initially starts up, the buzzer alarm may sound. This is normal. You can push the  
buzzer alarm button to cancel the sound. The buzzer alarm continues to work until the upper bottle  
An electrical power plug with a ground  
prong must be used to power the unit.  
This is to prevent electrical shock.  
vibration.  
temperature reaches the range of 2 to 8  
.
2)A place that is far away from any heat  
source  
3.Make sure that both sensor bottles have glycerin solution of 10% concentration. Please refer to  
item 4 in the Operation section for details.  
Do not use water lines to replace a  
properly installed ground wire for  
ground protection. This is because  
many water lines are actually built with  
non-conductors such as plastic.  
Avoid placing the unit near any heat dissipating  
devices such as a gas stove, radiator, oven and  
other source of heat. The refrigerator would lose  
its efficiency once the insulation is exposed to  
heat.  
4.This refrigerator has been factory preset to operate at a range of 2-8  
adjust the temperature set point.  
1. It is not necessary to  
5.It usually takes several hours for the unit to reach the preset temperature. Once the unit's  
temperature is stable, check the sensor bottle temperature to see if it matches the set point.  
6.Turn on the light for the interior. Inspect the light to makes sure it functions properly.  
7.Once a thorough inspection of the unit is completed, products can be loaded into the refrigerator.  
Beware to not load too many warm products into the unit at one time.  
Never use gas lines as the ground  
3)Do not install a refrigerator in direct sunlight  
If a unit is in direct sunlight, it may not function  
properly and its life expectancy can be shortened.  
protection for the refrigerator.  
action can be very dangerous.  
This  
Remote alarm contact  
Never use a telephone line or lightening  
rod as a grounding protection for the  
refrigerator. This is because during  
The terminals for the remote alarm contact are located on the back of the unit cover panel. The contacts  
are designed for 30 VDC and 2A load. The remote alarm contacts work in synchronization with the  
buzzer alarm on the refrigerator. Therefore, the remote alarm can be terminated by pushing the buzzer  
alarm button. However, in case of a power outage, the remote alarm contacts cannot be controlled by  
the buzzer alarm.  
4)Dry area  
Avoid placing the unit near damp areas such as a  
water faucet and sink.  
lightening, there is  
a
strong current  
present which is extremely dangerous.  
5)Clean area  
Avoid installing  
a
unit in or near chemical  
Remote alarm condition: Open for Normal, Close for Abnormal  
materials and materials that might have out-  
gassing property. Also avoid dusty locations.  
8)Do not place any objects on top of the  
refrigerator.  
6)Good ventilation  
There must be sufficient space around the unit for  
air ventilation. Lack of such space will reduce the  
unit's cooling capability.  
7)Ground protection  
Grounding the refrigerator will prevent electrical  
shock to operators if its electrical insulation  
becomes weak.  
Do not place the unit in an area where  
objects can fall directly onto it. This is to  
avoid damaging the refrigerator.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
Installation  
1)Supplementary power supply (9 volt DC).If the LED light is off and the recorder stops working, it  
would suggest that there might be issues associated with the main power supply. Please inspect  
whether there is a loss of power, a short circuit, or a loose connection. If there is no issue with the  
main power supply and the recorder starts working normally, please install the backup battery to the  
recorder. At this time, if the LED stops flashing, the supplementary power supply is normal. If the  
LED continues to flashes, it means that the power supply is under capacity requirement. It is  
necessary to replace the battery. The LED green light is off when there is sufficient power to the  
recorder.  
1)Remove packaging materials, packaging bag and straps. Open the refrigerator door to let it air out. If  
there are dirt marks on the outside of the unit, use a soft sponge with a light detergent solution to wipe  
the unit clean. Then, rinse the unit with a clean, slightly damp towel.  
2)Adjust the height of the unit. Turn the leveling leg screw counterclockwise to level the front end of the  
unit. The front casters should be off the ground.  
2)Remove the supplementary power supply. Power LED light continues to flash. The recorder  
works normally. It suggests that the main power supply is sufficient while the backup battery lacks  
capacity. It is then necessary to replace the battery until the green LED lights up and stops flashing.  
3.Changing of Recorder Chart Paper  
3)Preparation of sensor bottle Before using the refrigerator, inspect the sensor bottle located at the  
back of the drawers and shelves to see if the liquid level is up to the 200 mark inside the bottle. If the  
bottle has no fluid or the fluid level is lower than the 200 mark, use the following steps to fill the bottles at  
the upper corner and the lower corner with solution of 10% glycerin concentration (or other equivalent  
liquid such as glycol solution).  
When changing the chart paper for the recorder, please use the following steps.  
a.Locate the #3 button at the front of the recorder.  
b.Press and hold the #3 button for about one second. The imprinting stylus of the recorder moves to  
the left side of the chart.  
c.When the stylus is completely off the chart, unscrew the center nut for the paper and remove the  
old chart paper off the recorder. Place a new chart paper on the recorder. Carefully match the day  
and time line on the chart with the reference mark on the recorder panel (there is a small groove on  
the front of the recorder as shown in the figure).  
a)Remove the top drawer, bottom drawer, shelves and baskets.  
b)Unscrew the fastener that positions the bottle, remove the bottle.  
d.Place the center nut over the chart paper and tighten it. Press the #3 button to move the stylus  
back to the chart to continue the temperature recording.  
c) Unscrew the bottle's cap that is used for the sensor mounting.  
d)Pour a solution of 10% glycerin concentration into the bottle(s) to the 200 mark line.  
e)Mount the bottle back to the refrigerator liner wall.  
e.Inspect the contact between the pen and the paper chart. Adjust the arm if necessary to make  
contact. (Note, do not damage the pen and arm. It might be easier to remove the arm and pen to  
make a small bending adjustment and then install the pen back on the recorder. Repeat the process  
until the recorder arm and pen work properly.)  
f)Install the bottle back to its original position. Place the temperature sensor back inside the  
temperature bottle.  
f.The accuracy of the temperature must be checked after changing the chart paper. The procedure  
is described as follows. Press and hold the #3 button until the pen moves to the outside of the chart  
paper. Then press the #3 button again; the pen should return to the chart paper area. When the  
pen reaches the outer most circle of the chart paper, it pauses for a moment. If the pen does not  
match the outer circle of the chart marks, you can use the #2 or the #1 button to move the pen to  
match the line. This must be done within 5 seconds after the pen pauses. If the calibration is  
incomplete, repeat the same process until the pen matches the outer circle line.  
4.Calibration of Temperature Recorder  
This recorder has been calibrated at the factory. It retains the temperature accuracy and calibration even  
when there is a power supply interruption. Should there be a need to recalibrate the recorder, please  
follow the procedures below.  
a)Let the blood bank run continuously until it reaches a stable condition. The recorder has  
registered temperature for at least two hours.  
b)Place a standard thermocouple in a 100 ml container filled with 10% glycerin solution. Place the  
container and the temperature recorder sensor in the refrigerator. Do not place the recorder sensor  
in contact with any liquid.  
c)After four hours of soak time at the stable refrigerator temperature, the container temperature  
should match the refrigerator temperature. Then compare the thermocouple reading with the  
temperature recorder reading. If there is a discrepancy, use the #1 or the #2 button on the recorder  
to move the pen on the chart to reflect the correct temperature. Please note the pen moves five  
seconds after the button is released.  
Note:that the recorder has been calibrated in the factory. Unless it is absolutely necessary,  
please do not change the setting.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
Operation  
Temperature Recorder (optional)  
Control Panel  
The blood bank refrigerator is equipped with a 6-inch temperature recorder for the chamber  
temperature. When the unit is powered up, the recorder starts working.  
1)Temperature Indicator  
This refrigerator is factory preset to run at 5  
.
The high temperature alarm is set at 10 , the low  
temperature alarm is set at 2  
.
1.Installation and Operation  
Table 1, Temperature Indication  
For proper operation of the temperature recorder, please follow the procedures below.  
a)Open the recorder door to access the recorder.  
Definition  
of display  
Operation  
Push Button  
Display  
b)Plug in the 9 volt battery located on the upper right corner of the recorder. This battery is a  
standby power.  
Average Temperature in  
Sensor Bottle  
Average  
Temperature  
Plug in the unit, power is on  
1
c)Install a new chart by following the diagram below.  
Upper sensor bottle  
temperature  
Upper sensor  
bottle  
temperature  
Push temperature display  
button  
2
9 volt battery  
Lower sensor bottle  
temperature  
Lower Sensor  
bottle  
temperature  
Push temperature display  
button  
3
Chart Adjustment Buttons  
Average Temperature in  
Sensor Bottle  
Average  
Temperature  
Push temperature display  
button  
4
Imprinting Stylus  
Repeat Step 2  
5
Note: The digital display indicates either the upper bottle temperature or the lower bottle temperature  
Date Reference Mark  
respectively. The displayed temperature does not always stay at 5  
represents the average temperature inside the refrigerator.  
.
The temperature sensor  
(Figure 6) Chart Adjustment Buttons  
Center Nut and Chain  
Chart Paper  
Defrost Cycle  
(Figure 5)  
This blood bank refrigerator has two automatic defrost cycles.  
1)Cycle defrost  
2.Power Supply  
To maintain a constant temperature inside the blood bank, the refrigeration system cycles on and off  
according to the temperature demand. When the refrigeration compressor cycles off, a small electric  
heater is energized to thaw the frost off the evaporator surface. The function does not have an obvious  
impact to the chamber temperature.  
The recorder normally uses AC power supply as the refrigerator operates. If the AC power supply  
malfunctions, the LED indicator flashes to alert a power supply failure. The recorder continues to  
operate on its standby battery to track the refrigerator temperature. The typical battery can sustain the  
chart operation for about 30 hours. Please be aware that a weak battery should be replaced to avoid  
corrosion in the connector buttons. The battery power can be saved by disconnecting the connector  
button when the unit is not in service. During normal usage, make sure that the battery is plugged in to  
power the recorder in case of an AC power outage. The LED indicator glows continuously until the main  
power supply is on and the standby battery is replaced. When the battery is low, the LED indicator  
flashes to indicate that the battery needs to be replaced.  
2)Forced defrost cycle  
When the environmental humidity is too high or the products loaded into the refrigerator emit much  
moisture, cycle defrost will not be sufficient to remove all frost and ice accumulated on the evaporator  
surface. The equipment will start the forced defrost cycle. Once the frost is thawed, the refrigerator will  
resume normal operation. During the defrost process, the temperature of a 200-ml blood bag would  
probably rise by 1~ 2  
.
Replacement of Recorder Battery  
When the green LED on the recorder flashes, there are two ways to inspect the unit.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
Operation  
Note: Remote alarm contact and buzzer alarm work together. When the refrigerator is in alarm mode,  
the remote alarm contact can be disconnected by pushing the buzzer button. If the unit is in the alarm  
mode due to loss of power, the remote alarm contact remains in the alarm mode. This condition persists  
until the power is resumed. All alarms will automatically stop once the problem is corrected. If the alarm  
condition is not corrected within 30 minutes, the buzzer alarm and remote alarm contact are activated  
again. In the case of losing power, a fully charged battery can support the alarm system for 72 hours.  
For a newly started refrigerator or a refrigerator that has been unplugged for a long time, you should  
make sure the battery is fully charged. This charging process can take up to two days. When the battery  
discharges, the remote alarm contact continues to function.  
Alarm, Safety, and Auto-diagnostic Features  
This unit is equipped with all features listed in Table 2. Additionally, it also has auto-diagnostic features.  
Table 2, Alarm and Safety Features  
Alarm Methods and  
Indication  
Refrigerator Behaviors  
Alarm or Safety  
Buzzer Alarm  
Either upper sensor or  
lower sensor reaches  
10 and warmer  
High  
Temperature  
Alarm light flashes  
Alarm light flashes  
Buzzer alarm sounds  
Buzzer alarm sounds  
Either upper sensor or  
lower sensor is below  
2
Low  
Temperature  
Procedure to Test the Alarm System  
Push the “Alarm Test” button once. The alarm buzzer sounds three consecutive times in 1 Hz frequency.  
At the same time, the alarm light also flashes three consecutive times and the remote alarm contact  
opens and closes for three times. The alarm system functions properly.  
Temperature display  
shows for 60 seconds,  
and stops showing  
for 120 seconds. The  
process repeats until  
power is on again.  
After the time of  
losing power exceeds  
72 hours, buzzer  
alarm sounds.  
The unit lost power or  
the power switch is off.  
Loss of power  
Door ajar  
This blood bank refrigerator should be cared for by dedicated personnel. The operational status of  
the unit should be inspected and recorded daily. If the temperature runs toward over-temperature  
condition, you should transfer the stored products to another refrigerator. The products are only  
allowed back in the original unit only when operating errors are corrected.  
After a 10-minute  
delay, alarm light  
flashes.  
After a 10-minute  
delay, buzzer alarm  
sounds.  
Outer door is slightly  
open or wide open.  
Alarm light flashes.  
Temperature display  
shows “E1”.  
Upper temperature  
sensor circuit shorts  
or opens.  
Buzzer alarm sounds.  
Buzzer alarm sounds.  
Trouble Shooting Guide  
Problem  
Solution  
Alarm light flashes.  
Temperature  
display shows  
“E2”.  
Lower temperature  
sensor circuit shorts  
or opens.  
1.Inspect fuse blocks.  
2.Check that the cord is securely plugged into the power outlet.  
Unit stops operating.  
3.Inspect for problems associated with control wires.  
Abnormal  
sensor  
performance  
4.Make sure the voltage is not too low.  
Alarm light flashes.  
Temperature  
display shows  
“E3”.  
Defrost sensor circuit  
shorts or opens.  
1.Inspect to see if the unit is overloaded or products with warm temperature  
have been loaded into the unit.  
Buzzer alarm sounds.  
Buzzer alarm sounds.  
2.Make sure there is an air gap between products.  
3.Check to make sure the unit is not in direct sunlight or subject to any heat  
Unit's refrigeration is  
radiation.  
Alarm light flashes.  
Temperature display  
shows “E4”.  
Over temperature  
sensor circuit shorts or  
opens.  
not effective.  
Temperature tends to  
operate out of range.  
4.Investigate to see if there have been frequent door openings.  
5.Investigate to see if the environment temperature is too high.  
6.Inspect air channels to see if they are blocked.  
Alarm light stays  
on. Display shows  
temperature  
Rechargeable battery  
has low voltage, or  
there is no circuit .  
7.Check the condenser air filter to see if it is clogged with dirt.  
1.Make sure the unit is placed on a level floor.  
normally.  
The unit is too noisy.  
2.Make sure any part of the unit does not touch a wall or any object.  
1.If warm blood products are just loaded to the unit, allow ample time for the  
temperature to recover. The alarm signal cancels when the temperature  
recovers to normal level.  
2.Inspect the door to make sure it is shut properly. Door ajar alarm can sound  
if the door is slightly opened.  
Alarm light flashes,  
alarm buzzer sounds.  
3.There might be insufficient power supply. Let it run for some time to recover.  
4.Check the temperature of the unit to see if it is in over-temperature condition.  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Great States Network Card 204 14 User Manual
Haier Washer XQS70 98 User Manual
Haier Water Dispenser WDNS045 User Manual
Harbor Freight Tools Speaker 95953 User Manual
Hasbro Games Hand Held Game User Manual
Heath Zenith Door 6168 User Manual
Honeywell Thermostat CT3200 User Manual
Honeywell Video Game Keyboard HJC5000 User Manual
Hotpoint Washer Dryer WT761 User Manual
Hypertec Carrying Case Z0102 User Manual