Safety Information
700G Series
Pressure Gauge
PN 4098606
November 2011
© 2011 Fluke Corporation. All rights reserved. Printed in USA.
All product names are trademarks of their respective companies.
Download from Www.Somanuals.com. All Manuals Search And Download.
•
•
•
Read all safety Information before you
use the Product.
Safety Information
A Warning identifies conditions and procedures that are
dangerous to the user. A Caution identifies conditions
and procedures that can cause damage to the Product or
the equipment under test.
Do not use the Product in damp or wet
environments.
Batteries must only be changed in an
area known to be non-hazardous.
Explosion hazard.
XWWarning
To prevent possible electrical shock, fire, or
personal injury:
•
Do not use the RS-232 interface in
hazardous areas.
•
Use the Product only as specified, or the
protection supplied by the Product can
be compromised.
WCaution
To avoid possible damage to Product or to
equipment under test:
•
•
The battery door must be closed and
locked before you operate the Product.
•
If the display reads “OL” the range limit
is exceeded and the pressure source
must immediately be removed.
Replace the batteries when the low
battery indicator () shows to prevent
incorrect measurements.
•
Do not exceed the maximum torque
allowed. Maximum torque allowed is
13,5 Nm = 10 ftlbs.
•
Do not use and disable the Product if it
is damaged.
1
Download from Www.Somanuals.com. All Manuals Search And Download.
700G Series
Safety Information
Hazard Location Information/Approvals
Ex Hazardous Areas
Special Conditions for Safe Use
Misuse
An Ex-hazardous area as used in this manual refers to an
area made hazardous by the potential presence of
flammable or explosive vapors. These areas are also
referred to as hazardous locations, see NFPA 70 Article
500.
If the Product is exposed to overpressure or sudden
physical shock (such as being dropped) examine it for
any damage that can cause a safety concern. If
necessary, return the Product for evaluation to Fluke.
Refer to the How to Contact Fluke section.
)
® LR110460
Class I, Div. 2, Groups A-D
WWarning
To prevent possible fire, or personal injury:
•
Do not use the Product with flammable
substances.
(
II 3 G Ex nA IIB T6
KEMA 06ATEX0014 X
Ta=–10 °C... +55 °C
•
The Product is intended for installation
only in locations providing adequate
protection against the entry of solid
foreign objects or water capable of
impairing safety.
2
Download from Www.Somanuals.com. All Manuals Search And Download.
Pressure Gauge
Safety Information
Symbols
Symbols used on the Product and in this manual are
explained in Table 1.
Table 1. Symbols
Symbol
Meaning
Symbol
Meaning
W
Risk of danger. Important information. See manual.
Hazardous voltage. Risk of electrical shock.
P
Conforms to European Union directives.
Conforms to relevant North American
Safety Standards.
X
)
Do not dispose of this product as
unsorted municipal waste. Go to Fluke’s
website for recycling information.
f
Pressure
~
Conforms to relevant Australian standards.
(
Conforms to ATEX requirements
3
Download from Www.Somanuals.com. All Manuals Search And Download.
700G Series
Safety Information
How to Contact Fluke
To contact Fluke, call one of the following telephone numbers:
•
•
•
•
•
•
•
Technical Support USA: 1-800-44-FLUKE (1-800-443-5853)
Calibration/Repair USA: 1-888-99-FLUKE (1-888-993-5853)
Canada: 1-800-36-FLUKE (1-800-363-5853)
Europe: +31 402-675-200
Japan: +81-3-6714-3114
Singapore: +65-738-5655
Anywhere in the world: +1-425-446-5500
To view, print, or download the latest manual supplement, visit http://us.fluke.com/usen/support/manuals.