Model 122A21
Helium Bleed, ICP® Dynamic Pressure Sensor
Installation and Operating Manual
For assistance with the operation of this product , contact the Pressure
Division of PCB Piezotronics, Inc.
Division toll-free 888-684-0015
24-hour SensorLineSM 716-684-0001
Fax 716-686-9129
Download from Www.Somanuals.com. All Manuals Search And Download.
Warranty, Service, Repair, and
Return Policies and Instructions
The information contained in this document supersedes all similar information that
may be found elsewhere in this manual.
Total Customer Satisfaction – PCB
Piezotronics guarantees Total Customer
Satisfaction. If, at any time, for any
reason, you are not completely satisfied
with any PCB product, PCB will repair,
replace, or exchange it at no charge. You
may also choose to have your purchase
price refunded in lieu of the repair,
replacement, or exchange of the product.
recommended as this helps build
confidence in measurement accuracy and
acquired data. Equipment calibration
cycles are typically established by the
users own quality regimen. When in
doubt about a calibration cycle, a good
“rule of thumb” is to recalibrate on an
annual basis. It is also good practice to
recalibrate after exposure to any severe
temperature extreme, shock, load, or
other environmental influence, or prior
to any critical test.
Service – Due to the sophisticated nature
of the sensors and associated
instrumentation provided by PCB
Piezotronics, user servicing or repair is
not recommended and, if attempted, may
void the factory warranty. Routine
maintenance, such as the cleaning of
electrical connectors, housings, and
mounting surfaces with solutions and
techniques that will not harm the
physical material of construction, is
acceptable. Caution should be observed
to insure that liquids are not permitted to
migrate into devices that are not
hermetically sealed. Such devices should
only be wiped with a dampened cloth
and never submerged or have liquids
poured upon them.
PCB Piezotronics maintains an ISO-
9001 certified metrology laboratory and
offers calibration services, which are
accredited by A2LA to ISO/IEC 17025,
with full traceablility to N.I.S.T. In
addition to the normally supplied
calibration, special testing is also
available, such as: sensitivity at elevated
or cryogenic temperatures, phase
response, extended high or low
frequency response, extended range, leak
testing, hydrostatic pressure testing, and
others. For information on standard
recalibration services or special testing,
contact your local PCB Piezotronics
distributor, sales representative, or
factory customer service representative.
Repair – In the event that equipment
becomes damaged or ceases to operate,
arrangements should be made to return
the equipment to PCB Piezotronics for
repair. User servicing or repair is not
recommended and, if attempted, may
void the factory warranty.
Returning Equipment – Following
these procedures will insure that your
returned materials are handled in the
most expedient manner. Before returning
any equipment to PCB Piezotronics,
contact your local distributor, sales
representative, or factory customer
service representative to obtain a Return
Calibration – Routine calibration of
sensors and associated instrumentation is
Download from Www.Somanuals.com. All Manuals Search And Download.
Materials
Authorization
(RMA)
PCB for a complete statement of our
warranty. Expendable items, such as
batteries and mounting hardware, are not
covered by warranty. Mechanical
damage to equipment due to improper
use is not covered by warranty.
Electronic circuitry failure caused by the
introduction of unregulated or improper
excitation power or electrostatic
discharge is not covered by warranty.
Number. This RMA number should be
clearly marked on the outside of all
package(s) and on the packing list(s)
accompanying the shipment. A detailed
account of the nature of the problem(s)
being experienced with the equipment
should also be included inside the
package(s) containing any returned
materials.
A Purchase Order, included with the
returned materials, will expedite the
turn-around of serviced equipment. It is
recommended to include authorization
on the Purchase Order for PCB to
proceed with any repairs, as long as they
do not exceed 50% of the replacement
cost of the returned item(s). PCB will
provide a price quotation or replacement
recommendation for any item whose
repair costs would exceed 50% of
replacement cost, or any item that is not
economically feasible to repair. For
routine calibration services, the Purchase
Order should include authorization to
proceed and return at current pricing,
which can be obtained from a factory
customer service representative.
Contact Information – International
customers should direct all inquiries to
their local distributor or sales office. A
complete list of distributors and offices
Customers within the United States may
contact their local sales representative or
a
factory
customer
service
representative. A complete list of sales
representatives can be found at
numbers for a factory customer service
representative,
in
the
division
responsible for this product, can be
found on the title page at the front of this
manual. Our ship to address and general
contact numbers are:
PCB Piezotronics, Inc.
Warranty – All equipment and repair
services provided by PCB Piezotronics,
Inc. are covered by a limited warranty
3425 Walden Ave.
Depew, NY 14043 USA
Toll-free: (800) 828-8840
24-hour SensorLineSM: (716) 684-0001
E-mail: [email protected]
against
defective
material
and
workmanship for a period of one year
from date of original purchase. Contact
DOCUMENT NUMBER: 21354
DOCUMENT REVISION: B
ECN: 17900
Download from Www.Somanuals.com. All Manuals Search And Download.
OPERATION MANUAL FOR
MODELS 122A, 122A21, 122A22, and 122A24
HELIUM BLEED PRESSURE SENSOR
1
1.0
INTRODUCTION
The series of sensors described in this guide is
designed to measure dynamic pressure in intense
thermal environments typified by rocket engine
combustion chambers.
These sensors feature acceleration-compensated
piezoelectric pressure probes, mounted in a conical
end seal helium bleed adaptor.
The helium gas, applied from a 1500 psi source, is
metered by a built-in orifice and flows around the
probe body and out the passage of the adaptor ahead
of the probe diaphragm. The helium gas flowing out
the passage protects the sensor diaphragm from the
hot gasses as well as increasing the natural frequency
of the passage by a factor of approximately three to
one.
2.0
DESCRIPTION
The model 122A contains a one pC/psi pressure
probe (sensor) and is designed to be used with
electrostatic charge amplifiers such as the PCB
Model 462A.
Series 122A: Helium Bleed Rocket Motor, ICP®
and Charge Sensors
The Models 122A21, A22, and A24 contain built-in
impedance converter electronics circuits which
convert the high-impedance voltage from the quartz
crystal to a low-impedance voltage signal that can be
fed directly into most common indicating or
recording instruments.
NOTE: The sensor probe in this series is factory
installed at a precise depth in the outer body. Do not
attempt removal of this probe. Contact the factory
should any questions or problems arise concerning
the sensor probe.
The 122A series sensors contain a 10 micron helium
filter contained in a fitting threaded onto the helium
input tube. The purpose of this filter is to prevent
particles in the helium supply from plugging the
metering orifice.
The Models 122A21, A22, and A23 Sensors use low
output-impedance probes (ICP) and will not require
the insulation resistance cautions described for Model
122A.
Connect the sensor cable to the appropriate ICP
power unit and check the front panel monitor meter to
check cable connections and internal amplifier.
Consult Guide G-0001B, "General Guide to ICP
Instrumentation" for installation and usage hints for
ICP instruments.
The outer body of this series is 17-4 PH hardened
stainless steel and features a conical tip for sealing
without seal rings or washers. This feature also
allows the sensor sensing passage to be placed closer
to the chamber without a long passage that could
cause resonance problems, severely limiting high-
frequency response.
Drawing number: 21089
Revision: NR
Download from Www.Somanuals.com. All Manuals Search And Download.
OPERATION MANUAL FOR
MODELS 122A, 122A21, 122A22, and 122A24
HELIUM BLEED PRESSURE SENSOR
2
3.0
INSTALLATION
4.0
4.1
OPERATION
Prepare mounting port in accordance with
instructions on installation drawing 122-1010-90.
CHARGE OUTPUT MODEL
122A
Inspect the mounting port for burrs and tool marks at
the conical seal surface.
For the 122A Charge Output Model, it is only
necessary to select the proper range on the charge
amplifier and proceed with the measurement.
NOTE: A 75 end-cutting reamer, Model 040A02, is
available from PCB. Contact the factory for price
and delivery information.
The probe sensitivity is approximately 1.0 pC/psi.
(The exact sensitivity is given on the calibration
certificate supplied with each instrument.)
Use recommended mounting torque when installing
sensors. Do not overtorque as damage to seal surface
could result.
EXAMPLE:
The dynamic pressure expected is
500 psi and the desired output voltage for this
pressure is 1 volt.
NOTE: When attaching helium supply hose, do not
torque hose fitting through helium tube on sensor.
Simply set the range switch of the charge amplifier to
500 units per volt and dial the standardization pot to
the exact sensor sensitivity at 500 psi (from
calibration graph). The system will then yield exactly
1 volt for 500 psi of input pressure.
Back up fitting with wrench on 3/8 hex on helium
hose fitting. (Refer to note 3 on installation drawing.)
In the presence of high vibration, support helium
supply tube by clamping to rigid surfaces with
adequate strain relief loop to avoid stressing helium
tube on sensor.
4.2
TIME CONSTANT
During in-place calibration, static methods (dead
weight tester) may be used in a drift-free system by
placing the charge amplifier in "LONG TC" position.
For charge output Model 122A, connect 4 ft. cable to
charge amplifier and switch to "long" time constant
position and check for drift.
However, during actual use, the TC mode switch
should be in "SHORT" position for two reasons.
Clean
offending
cable
connections
with
Trichloroethylene or equivalent solvent to restore
insulation resistance.
Often times, the dynamic pressure being measured
exists superimposed upon a higher static level. The
short TC position will allow the effect of this static
level to zero out quickly without effect on the
dynamic level indicated.
It is good practice to depress ground button (or place
switch in GND" position) on the charge amplifier
while attaching input cables to the amplifier. This
will prevent spurious charges which may accumulate
on sensors and cables from blowing out the input
circuitry of the charge amplifier.
Also, the influx of large amounts of thermal energy,
as in rocket combustion measurements, can cause the
sensor to generate spurious transient outputs of
considerable magnitude. Using the charge amplifier
on SHORT TC alleviates this problem by shunting
these longer term charge outputs to ground while
measuring the faster dynamic pressure pulses.
Since the charge output sensors are susceptible to
decreased insulation resistance, use care to keep cable
connections clean by use of shrink tubing wherever
possible.
Drawing number: 21089
Revision: NR
Download from Www.Somanuals.com. All Manuals Search And Download.
OPERATION MANUAL FOR
MODELS 122A, 122A21, 122A22, and 122A24
HELIUM BLEED PRESSURE SENSOR
3
OPERATION - ICP MODELS
5.1
CHARGE MODEL 122A
4.3
122A21, A22, and A24
To calibrate this model, install in pressure port of
dead weight tester and apply various static pressures
recording corresponding outputs at each pressure of
interest.
Consult Guide G-0001B for a complete description of
the low-impedance concept in instrumentation.
All PCB power units designed for ICP use contain
built-in fault monitor meters to aid in locating circuit
faults (shorts or opens) and to identify normal
operation of the miniature impedance converting
amplifier built into the sensor.
Discharge the charge amplifier before each reading
and return pressure to zero before each new pressure
input.
Use the charge amplifier on "LONG" time constant
and keep cables and connections clean to avoid drift.
Do not attempt to apply voltage directly to the input
pin of the sensor from any other power source. The
input current must be limited to 20 mA maximum and
the input voltage must not be higher than 35 VDC.
Consult the factory if it is desired to build a power
unit rather than to use one of those available from
PCB.
A D.C. digital voltmeter is ideal for readout during
calibration.
NOTE: The helium tube must be plugged with a
pressure-tight seal before attempting to calibrate to
avoid losing pressure through the helium tube.
The Models 122A21, 122A22 and 122A24 are
identical except for sensitivity.
If a hydraulic calibration system is used, be sure to
remove all traces of oil from the diaphragm passage.
The Model 122A21 has a sensitivity of 20mV/psi
which gives a full-scale range (for +5 V out) of 250
psi.
5.2
ICP MODELS 122A21, 122A22
and 122A24
The Model 122A22 has a sensitivity of 1.0 mV/psi
and a full-scale range of 5000 psi.
Because of the shorter time constants built into these
models, static calibration methods are much more
difficult to use.
The Model 122A24 has a sensitivity of 5.0 mV/psi
and a full-scale range of 1000 psi.
Pressures must be applied rapidly and readout must
be made with fast recording devices such as light
storage oscilloscopes or similar types of data
acquisition hardware.
NOTE: Full-scale range is limited by the helium
source pressure. The helium source pressure must be
at least twice the peak combustion pressure (static
plus dynamic) to be measured, but not more than
about ten times the lowest pressure.
A pneumatic (compressed air or hydrogen) system
with quick opening valve and reference gage is an
excellent method of calibrating these sensors.
This is true for all models, charge or ICP.
To take full advantage of the time constant built into
the sensor, use a DC-coupled power unit such as the
PCB Model 484B06.
5.0
CALIBRATION
Drawing number: 21089
Revision: NR
Download from Www.Somanuals.com. All Manuals Search And Download.
OPERATION MANUAL FOR
MODELS 122A, 122A21, 122A22, and 122A24
HELIUM BLEED PRESSURE SENSOR
4
6.0
GROUND ISOLATION
All sensor probes in this series are mounted in an
"OFF GROUND" configuration. That is, the probe
body is electrically insulated from the sensor outer
body to avoid ground loops.
7.0
MAINTENANCE AND REPAIR
The sealed construction and miniature size of these
sensors make field repair impractical.
The helium filter element should be cleaned should
the flow rate decrease. Remove the element and
ultrasonically agitate in solvent for 15 minutes, then
dry.
Should the orifice become plugged, back pressure the
orifice by applying air pressure to diaphragm end of
sensor with filter off and helium tube vented to
atmosphere.
The only maintenance possible on the probe is
maintaining of insulation resistance of the Model
122A probe/cable assembly. Clean connectors with
suitable solvent and dry.
Contact factory for assistance should problems
arise.
ICP is a registered trademark of PCB Piezotronics
Drawing number: 21089
Revision: NR
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
OPERATION MANUAL FOR
MODELS 122A, 122A21, 122A22, and 122A24
HELIUM BLEED PRESSURE SENSOR
7
Test Report
Helium Flow Test and Burst Test
Sensor Model ________________________
S/N __________________________
Initial Calibration:
Input (PSI)
System Output (V)
500
1000
1000
1500
1500
2000
2000
2500
2500
3000
3000
Burst Test to 5000 PSI - Failed
Final Calibration:
Input (PSI)
500
System Output (V)
Helium Flow Test:
Input Pressure
Flow Rate
_________ PSIG
_________ SCFH
Date: _____________________
By: _____________________
Drawing number: 21089
Revision: NR
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
|