Coronis Fusion 4MP/6MP
User Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Welcome!
Thank you for choosing this CORONIS FUSION 4MP/6MP display!
Coronis Fusion is an ingenious PACS display system designed to further
enhance flexibility and productivity in diagnostic imaging.
Featuring the industry’s first 30-inch color LCD that can be used as two
seamless heads or one wide-screen display, Coronis Fusion offers you
the freedom to organize your workspace just the way you want it. It
allows you to read CT, MR, cath and echo cardiogram images, or any
other combination, side by side on a single diagnostic screen.
Coronis Fusion features the latest, breakthrough In-Plane Switching (IPS-
Pro) LCD technology. This technology brings the most advanced LCD
viewing characteristics on the market, setting new standards for
brightness and contrast, even from a wide viewing angle.
Use the instructions in this guide to install your CORONIS FUSION 4MP/6MP
display and discover all these interesting features yourself!
Following symbols may be used throughout this guide:
Warning: Risk of
injury to human
beings
Caution: Risk of
damage to the
product
WARNING
Important notice
or remark
Note
Hint, tip
Additional
information
operating your CORONIS FUSION 4MP/6MP display.
Download from Www.Somanuals.com. All Manuals Search And Download.
Coronis Fusion 4MP/6MP
3
Download from Www.Somanuals.com. All Manuals Search And Download.
Coronis Fusion 4MP/6MP
4
Table of contents
Routing the cables................................................................................17
First time starting up............................................................................21
Info messages.......................................................................................27
Download from Www.Somanuals.com. All Manuals Search And Download.
Coronis Fusion 4MP/6MP
5
Front glass .............................................................................................49
Cabinet...................................................................................................50
Safety information................................................................................51
EMC notice.............................................................................................59
Legal disclaimer....................................................................................66
Download from Www.Somanuals.com. All Manuals Search And Download.
Coronis Fusion 4MP/6MP
6
What's in the box
Your CORONIS FUSION 4MP/6MP display comes with:
• this CORONIS FUSION 4MP/6MP display guide
• a system CD
• two DVI cables
• a USB cable
• a set of AC power cords
• an external power supply
If you ordered a Barco display controller, it’s also in the box together
with its accessories. A dedicated user guide is available on the system
CD.
Keep your original packaging. It is designed for this display and
is the ideal protection during transport.
Download from Www.Somanuals.com. All Manuals Search And Download.
Coronis Fusion 4MP/6MP
7
Parts, controls and connectors
Front view
1
2
3
4
5 6 7
8
1
3
5
7
2
4
6
8
USB downstream connectors
Right/Up key
Left/Down key
Enter key
Stand-by key
Ambient light sensor
Control wheel
Power status indicator light
Download from Www.Somanuals.com. All Manuals Search And Download.
Coronis Fusion 4MP/6MP
8
CORONIS FUSION 4MP
1
2
3
4
5
6
7
1
3
2
4
+24 VDC power input
USB upstream connector
USB downstream connector
+5 VDC, 0.25 A power output
for accessories
5
7
6
+3.3 VDC, 0.25A auxiliary
power output
DVI 1 video input
DVI 2 video input
Download from Www.Somanuals.com. All Manuals Search And Download.
10 Coronis Fusion 4MP/6MP
CORONIS FUSION 6MP
1
2
3
4
5
6
7
1
3
2
4
+24 VDC power input
USB upstream connector
USB downstream connector
+5 VDC, 0.25 A power output
for accessories
5
7
6
+3.3 VDC, 0.25A auxiliary
power output
DVI 1 video input
DVI 2 video input
Download from Www.Somanuals.com. All Manuals Search And Download.
Coronis Fusion 4MP/6MP 11
Display installation
Prior to installing your CORONIS FUSION 4MP/6MP display and
connecting all necessary cables, make sure to have a suitable
display controller physically installed in your computer. If you are
using a Barco display controller, please consult the dedicated user
guide available on the system CD.
Removing the covers
CONNECTOR COMPARTMENT COVER
The connector compartment cover should be removed to get access to
the connectors. To remove the cover, proceed as follows:
1 Gently lift the clips at one of the handles of the connector
compartment cover to release that side of the cover.
2 Do the same at the other side of the cover.
3 Remove the cover.
Download from Www.Somanuals.com. All Manuals Search And Download.
12 Coronis Fusion 4MP/6MP
DISPLAY STAND COVER
To get access to the cable routing channel, the display stand cover
should be removed. This can be done as follows:
1 Gently press and hold the clips at the inside top of the cover.
2 Slide the cover upwards while holding the clips pressed.
Download from Www.Somanuals.com. All Manuals Search And Download.
Coronis Fusion 4MP/6MP 13
Unlocking the height mechanism
In the factory, the height-positioning system in the display stand is
locked with a red hook to prevent damage during transportation. You’ll
have to remove this hook before adjusting your display height position.
1 While holding the display panel pushed down, pull out the red
hook in the display stand.
To retain the hook for possible future transportation, insert the
short, red end of the hook back into the stand of your display.
Download from Www.Somanuals.com. All Manuals Search And Download.
14 Coronis Fusion 4MP/6MP
Adjusting the display position
Now that the height-positioning system of your display is unlocked, you
can tilt, swivel, raise and lower the display as desired.
Do not try to pivot your display when attached to the stand.
Trying to do so could cause serious damage to your display and
its stand.
Download from Www.Somanuals.com. All Manuals Search And Download.
Coronis Fusion 4MP/6MP 15
Connecting the signal cables
1 Connect one head of the display controller to the DVI 1 connector
with one of the supplied DVI cables.
2 Connect another head of the display controller to the DVI 2
connector with one of the supplied DVI cables.
3 If you want to make use of your display’s USB downstream
connectors, connect a PC USB downstream connector to the
display’s USB upstream connector by means of the supplied USB
2.0 cable.
Download from Www.Somanuals.com. All Manuals Search And Download.
16 Coronis Fusion 4MP/6MP
Connecting the power cables
1 Connect the supplied external DC power supply to the +24 VDC
power input of your CORONIS FUSION 4MP/6MP display.
2 Plug the other end of the external DC power supply into a
grounded power outlet by means of the proper power cord
delivered in the packaging.
Routing the cables
Route all connected cables trough the cable routing channel in the stand
of your display.
The cable straps at the inside of the connector compartment
allow you to fix the cables for better shielding of the cables.
Download from Www.Somanuals.com. All Manuals Search And Download.
Coronis Fusion 4MP/6MP 17
Re-attaching the covers
1 Re-attach the display stand cover by sliding it downward. You’ll
hear a “click” sound of the cover’s clips when the display stand
cover is in position.
Pay attention that all cables stay in the cable channel
while replacing the cover.
2 Re-attach the connector compartment cover by sliding the cover’s
top in position and then push the cover’s bottom. You’ll hear a
“click” sound of the cover’s clips when the connector compartment
cover is in position.
Download from Www.Somanuals.com. All Manuals Search And Download.
18 Coronis Fusion 4MP/6MP
VESA-mount installation
The panel, standard attached to a stand, is compatible with the VESA
100 mm standard. Thus, it can be used with an arm that is approved by
VESA. This chapter shows you how to release the panel from the stand
and how to attach it to an arm. If you’re not using an arm, you can skip
this chapter.
1 Gently pull open both lower sides of the cover.
2 Slide the cover upwards while holding the lower sides pulled open.
Download from Www.Somanuals.com. All Manuals Search And Download.
Coronis Fusion 4MP/6MP 19
3 Lift the plastic frame that covers the fixation of the panel to the
stand. Turn it for 45 degrees to uncover the fixation screws.
4 Unscrew the four fixation screws fixing the panel to the stand.
5 Attach the panel firmly to the arm using 4 screws M4 x 10 mm.
You should mount the panel in landscape position.
Portrait position is possible but not supported.
Use an arm that can support a weight of at least 21.6 kg
(47.62 lbs). Failure to do so could make the panel fall,
causing serious injury to a child or adult, and serious
damage to the equipment.
Never move a display attached to an arm by pulling or
pushing the display itself. Instead, make sure that the
arm is equipped with a VESA approved handle and use
this to move the display.
Please refer to the instruction manual of the arm for more
information and instructions.
Download from Www.Somanuals.com. All Manuals Search And Download.
20 Coronis Fusion 4MP/6MP
First time starting up
You are now ready to start up your CORONIS FUSION 4MP/6MP display
system for the first time.
1 Switch on your CORONIS FUSION 4MP/6MP display as described in
2 Turn on the computer connected to your display.
If you have properly installed your display and display controller, the
Windows start-up messages will appear once the boot procedure is
finished.
Your CORONIS FUSION 4MP/6MP display will be running in a basic
video mode at a default refresh rate when first time starting
up.
If you are using a Barco display controller, please consult the
dedicated user guide available on the system CD to install the
drivers, software and documentation. When this is done, your
display will automatically detect the connected video input
signal(s) and apply the correct video mode and refresh rate.
Download from Www.Somanuals.com. All Manuals Search And Download.
Coronis Fusion 4MP/6MP 21
Daily operation
Recommendations for daily operation
OPTIMIZE THE LIFETIME OF YOUR DISPLAY
Enabling the Display Power Management System (DPMS) of your display
will optimize its diagnostic lifetime by automatically switching off the
backlight when the display is not used for a specified period of time. By
default, DPMS is enabled on your display, but it also needs to be
activated on your workstation. To do this, go to “Power Options
Properties” in the “Control Panel”.
Barco recommends setting DPMS activation after 20 minutes of
non-usage.
USE A SCREEN SAVER TO AVOID IMAGE RETENTION
Prolonged operation of an LCD with the same content on the same
screen area may result in a form of image retention.
You can avoid or significantly reduce the occurrence of this phenomenon
by using a screen saver. You can activate a screen saver in the “Display
properties” window of your workstation.
Barco recommends setting screen saver activation after 5
minutes of non-usage. A good screen saver displays moving
content.
In case you are working with the same image or an application with
static image elements for several hours continuously (so that the screen
saver is not activated), change the image content regularly to avoid
image retention of the static elements.
UNDERSTAND PIXEL TECHNOLOGY
LCD displays use technology based on pixels. As a normal tolerance in
the manufacturing of the LCD, a limited number of these pixels may
remain either dark or permanently lit, without affecting the diagnostic
Download from Www.Somanuals.com. All Manuals Search And Download.
22 Coronis Fusion 4MP/6MP
performance of the product. To ensure optimal product quality, Barco
applies strict selection criteria for its LCD panels.
To learn more about LCD technology and missing pixels, consult
medical.
ENHANCE USER COMFORT
Every Barco multi-head display system is color matched with the highest
specification in the market.
Barco recommends keeping color-matched displays together.
Furthermore, it is important to use all displays of a multi-head
configuration at the same rate to preserve color matching
throughout the economic lifetime of the system.
MAXIMIZE QUALITY ASSURANCE
The ‘MediCal QAWeb’ system offers online service for high-grade Quality
Assurance, providing maximum diagnostic confidence and uptime.
Barco recommends to install MediCal QAWeb Agent and apply
the default QAWeb policy at least. This policy includes
calibration on regular intervals. Upgrading to MediCal QAWeb
Server offers even more possibilities.
Learn more and sign up for the free MediCal QAWeb Essential
Download from Www.Somanuals.com. All Manuals Search And Download.
Coronis Fusion 4MP/6MP 23
Stand-by switching
The connected power supply also provides a switch that can be
used to turn the power completely off. To use the display,
please make sure to switch on this power supply. This can be
done by pushing the on/off switch on the power supply into
the "|" position.
USING THE TOUCH KEYS:
1 While the OSD is not on the screen, touch any touch key. As a
result, the control panel illumination is switched on for a few
seconds.
2 While the contol panel illumination is on, touch the stand-by
touch key. A message appears on the screen.
3 Touch the stand-by touch key once more to confirm.
4 To switch the display on while it is in stand-by, touch any touch
key. Next, while the contol panel illumination is on, touch the
stand-by touch key.
USING THE CONTROL WHEEL:
1 While the OSD is not on the screen, click the control wheel shortly.
A message appears on the screen.
2 Click the control wheel once more to confirm.
3 To switch the display on while it is in stand-by, click the control
wheel shortly.
Download from Www.Somanuals.com. All Manuals Search And Download.
24 Coronis Fusion 4MP/6MP
Bringing up the OSD menu
USING THE TOUCH KEYS:
1 While the OSD is not on the screen, touch any touch key. As a
result, the control panel illumination is switched on for a few
seconds.
2 While the contol panel illumination is on, touch the Enter
touch
key.
USING THE CONTROL WHEEL:
While the OSD is not on the screen, rotate the control wheel.
Navigating through the OSD menus
USING THE TOUCH KEYS:
1 Use the Up and Down keys to move through the menus.
2 To go into a submenu, use the Up or Down key to highlight
the submenu. Next, touch the Enter
submenu.
key to go into the
3 When you select Exit, the OSD exits the submenu and jumps to the
previous (higher-level) menu. When you select Exit in the main
menu, the OSD is hidden.
USING THE CONTROL WHEEL:
1 Rotate the control wheel to move up and down through the
menus.
2 To go into a submenu, rotate the control wheel to highlight the
submenu. Next, click the control wheel to go into the submenu.
3 When you click on Exit, the OSD exits the submenu and jumps to
the previous (higher-level) menu. When you click on Exit in the
main menu, the OSD is hidden.
Download from Www.Somanuals.com. All Manuals Search And Download.
Coronis Fusion 4MP/6MP 25
Making selections in the OSD menus
USING THE TOUCH KEYS:
1 Navigate through the menus and highlight the parameter you wish
to select.
2 Touch the Enter
key to make a selection or toggle between
several settings.
USING THE CONTROL WHEEL:
1 Navigate through the menus and highlight the parameter you wish
to select.
2 Click the control wheel to make a selection or to toggle between
several settings.
Changing values in the OSD menus
USING THE TOUCH KEYS:
1 Navigate through the menus and highlight the parameter you wish
to adjust.
2 Touch the Enter
key to go into edit mode.
3 Use the Up or Down key to change the value.
4 Touch the Enter
key to enter the new value and return to the
menu.
USING THE CONTROL WHEEL:
1 Navigate through the menus and highlight the parameter you wish
to adjust.
2 Click the control wheel to go into edit mode.
3 Rotate the wheel to change the value.
4 Click the wheel to enter the new value and return to the menu.
Download from Www.Somanuals.com. All Manuals Search And Download.
26 Coronis Fusion 4MP/6MP
Advanced operation
Info messages
The following messages may appear automatically on the screen:
INPUT SELECTION
The Input selection menu appears in case 2 video signals are connected
and the resolution of one or both video signals is too high to display
them both.
In that case the input selection menu allows you to select one of both
video signals.
NO SIGNAL SELECTED
This message appears in case the display finds no proper video signal at
its inputs.
RESOLUTION TOO HIGH
This message appears in case the resolution of the selected video signal
is too high to display.
TEXT MODE / DIAGNOSTIC MODE
These messages appear for a few seconds directly after switching to Text
mode or Diagnostic mode.
In text mode, the display luminance is approximately half of the
luminance in diagnostic mode. Text mode is intended for using the
display with office applications such as word processing. You can switch
between both modes by touching the up and down keys at the
same time.
Download from Www.Somanuals.com. All Manuals Search And Download.
Coronis Fusion 4MP/6MP 27
LCM TEMPERATURE TOO HIGH
This is a serious error message. You should power off the
display and call for service.
The warning means the LCD panel temperature is too high. As a result
the display switches to a safe operating mode (with decreased backlight
value). The message will not disappear until the display is powered off.
LCM TEMPERATURE SENSOR OFF LINE
This is a serious error message. You should power off the
display and call for service.
The warning means the LCD panel temperature sensor has lost
communication. As a result the display switches to a safe operating
mode (with decreased backlight value). The message will not disappear
until the display is powered off.
Download from Www.Somanuals.com. All Manuals Search And Download.
28 Coronis Fusion 4MP/6MP
On-screen display (OSD)
Luminance and color menu
Measured luminance
Indicates the actual luminance measured
by the internal sensor. This is a read-only
value. It is expressed in cd/m2.
Luminance target
Color target
Allows to manually adjust the luminance
target. See note below.
Allows to select from a list of factory-
defined and user-defined color targets.
Viewing mode
Allows to select the viewing mode:
Diagnostic or Text. In text mode the
luminance is approximately half of the
luminance in diagnostic mode. This is
intended for use with office applications
such as word processing. You can also
switch between both modes by touching
the up and down keys at the same
time while the OSD is not on the screen.
Color definition
Jumps to the color definition submenu,
which allows to change the definition of
the user-defined color targets.
When you change the luminance target, the display will adjust
its backlight to reach the target. This can be seen in the
Measured luminance line.
When the luminance target cannot be reached, e.g., due to
aging of the backlight, the Measured luminance line changes
to Minimum value reached or Maximum value reached.
Download from Www.Somanuals.com. All Manuals Search And Download.
Coronis Fusion 4MP/6MP 29
COLOR DEFINITION SUBMENU
A color target is defined by its luminance target and color temperature.
The color definition submenu allows to edit these values.
Color name
Indicates the color target you are editing.
You can select another color target to edit
by using the Up and Down touch keys
or by rotating the control wheel.
Luminance target
Defined in
Allows to select the luminance target that
corresponds to the color target.
Indicates how the color temperature is
defined: in (x,y) co-ordinates or Kelvin (K).
You can switch between both by using the
Up and Down touch keys or by
rotating the control wheel.
This function is available only when the
User color target is selected.
x
Allows to define the x co-ordinate of the
color temperature that corresponds to the
color target.
This function is available only when the
color target is defined in (x,y) co-ordinates.
This function is not available when the
Native White color target is selected.
y
Allows to define the y co-ordinate of the
color temperature that corresponds to the
color target.
This function is available only when the
color target is defined in (x,y) co-ordinates.
This function is not available when the
Native White color target is selected.
Download from Www.Somanuals.com. All Manuals Search And Download.
30 Coronis Fusion 4MP/6MP
Color temperature
Allows to define the color temperature that
corresponds to the color target in Kelvin
units.
This function is available only when the
color target is defined in Kelvin units.
This function is available only when the
User color target is selected.
Restore factory x, y
values
Allows to reset the (x,y) co-ordinates of
the color temperature to the factory values.
This function is not available when the
Native White color target is selected.
Display Function menu
Display function
Allows to select from a list of pre-defined
display functions.
If the DICOM DF is selected, a number of
additional settings is available.
ALC & DICOM options
Jumps to the ALC & DICOM Options
submenu, which allows to edit the settings
for the DICOM display function.
This function is available only when the
DICOM DF is selected.
Download from Www.Somanuals.com. All Manuals Search And Download.
Coronis Fusion 4MP/6MP 31
ALC & DICOM OPTIONS SUBMENU
Measured ambient
light
Indicates the ambient light actually
measured by the ambient light sensor on
the front of the display. This is a read-only
value.
Averaged ambient
light
Shows the average of the measured
ambient light since the display was
switched on.
When you execute the function “Measure
ALC Value”, this value will be stored as the
Measured ALC correction value for the
selected reading room, unless this value is
higher than the maximum ambient light
determined for the selected reading room.
Correction value
Cont. DICOM ALC
Shows the ambient light correction value
that is taken into account in the calculation
of the display function.
Allows to switch continuous DICOM ALC
on/off.
When switched on, the DICOM DF is
recalculated continuously taking the
averaged ambient light into account.
When switched off, the DICOM DF is
recalculated only at the moment the
DICOM Offset is changed (see below).
Download from Www.Somanuals.com. All Manuals Search And Download.
32 Coronis Fusion 4MP/6MP
DICOM offset
This function is available only when the
DICOM DF is selected.When Continuous
DICOM ALC is switched off, the DICOM DF is
recalculated at the moment the DICOM
Offset is changed, taking the new DICOM
Offset into account.
• When set to “Dark Room”, the
ambient light is not taken into
account.
• When set to “Preset”, a preset
ambient light value determined by
the selected reading room is taken
into account.
• When set to “Measured ALC”, the
averaged ambient light, determined
by the selected reading room, is
taken into account.
This function is available only when
Continuous DICOM ALC is off.
Reading room
Allows to select from a pre-defined list of
reading room types.
You must select a reading room that
corresponds to the type of room the
display is installed in.
This function is available only when
Continuous DICOM ALC is off.
Reading room def.
Jumps to the reading room definition
submenu, which allows to edit the reading
room condition settings.
This function is available only when
Continuous DICOM ALC is off.
Download from Www.Somanuals.com. All Manuals Search And Download.
Coronis Fusion 4MP/6MP 33
Calibration info
Jumps to the calculation information
submenu, which displays information
about the values taken into account to
recalculate the DICOM DF.
READING ROOM DEFINITION SUBMENU
Measured ambient
light
Indicates the ambient light actually
measured by the ambient light sensor on
the front of the display. This is a read-only
value.
Averaged ambient
light
Shows the average of the measured
ambient light since the display was
switched on. This is a read-only value.
Reading room
Indicates the reading room type you are
editing. You can select another room to
edit by using the Up or Down touch
keys or rotating the control wheel.
Max. ambient light
This indicates the maximum ambient light
that corresponds to the selected reading
room.
If the measured ambient light is higher
than the value entered here, you should
take measures to darken the room or
select another reading room type.
Preset corr. value
This value is taken into account in the
calculation of the DICOM DF if Continuous
DICOM ALC is switched off and DICOM
Offset is set to “Preset”.
To each reading room type corresponds
another preset value.
Download from Www.Somanuals.com. All Manuals Search And Download.
34 Coronis Fusion 4MP/6MP
Measured ALC value
The value shown here is taken into account
in the calculation of the DICOM DF if
Continuous DICOM ALC is switched off and
DICOM Offset is set to “Measured ALC”.
When you select this line and confirm the
selection, the actual averaged ambient
light value is stored here, overwriting the
former value.
However, if the averaged ambient light
value is higher than the maximum ambient
light value that corresponds to the selected
reading room, a warning is displayed
prompting you to dim the light in the
room, and the value will be limited to the
maximum ambient light value.
If you dim the light in the room you can
execute the Measure ALC Value function
again to enter the new, lower value.
CALIBRATION INFORMATION SUBMENU
Preset ambient value Shows the ambient light correction value
/ Measured ALC
ambient value /
Average ambient
value
taken into account to calculate the DICOM
DF, expressed in lux.
Bright luminance
Shows the bright luminance value taken
into account to calculate the DICOM DF.
Dark luminance
Shows the dark luminance value taken into
account to calculate the DICOM DF.
Ambient correction
Shows the ambient light correction value
taken into account to calculate the DICOM
DF, expressed in cd/m².
Download from Www.Somanuals.com. All Manuals Search And Download.
Coronis Fusion 4MP/6MP 35
Settings menu
DPMS
Allows to switch the display power
management system on/off. See note
below.
Power LED
Allows to switch the power LED’s on state
on/off. The LED's orange DPMS state is not
influenced by this setting. So, when the
display goes into power-saving mode, the
LED will turn orange, even if it was
switched off by this setting
User controls
Allows to disable the user controls on the
front.
When switched off, the user cannot display
the OSD until the user controls keycode is
entered.
To enable the user controls again, To
Sound
Allows to switch the sound on/off. When
switched on, a short beep sounds each
time you touch a soft touch key.
Auto menu exit
Allows to switch the automatic menu exit
on/off. When switched on, the OSD is
closed automatically when left idle for a
certain time.
Download from Www.Somanuals.com. All Manuals Search And Download.
36 Coronis Fusion 4MP/6MP
OSD Orientation
Select the desired orientation of the on-
screen display. The default position is
Landscape, but if you put the display in
portrait orientation (possible but not
supported), you should select Portrait as
OSD orientation as well. If not, the OSD will
not be easily readable.
Input Switch Mode
Select “Auto” if you wish the display to
select the video inputs automatically. If
only one video input is connected, the
image will be shown in the center of the
screen. If both video inputs are connected,
the video signal connected to the “Left”
input is shown on the left side of the
screen and the signal connected to the
“Right” connector is shown on the right
side. If one of the video signals exceeds
the resolution of 1640x2048, the signal is
suppressed.
Select “Left” to display only the signal
connected to the “Left” video input.
Select “Right” to display only the signal
connected to the “Right” video input.
Preferred Input
Select the video resolution you wish the
connected graphic board(s) to reproduce.
Before changing this setting you must
switch off the PC. After changing the
setting, switch on the PC so that the
graphic board changes its resolution.
This function takes effect with plug-and-
play graphic boards only.
More information can also be found in the
Getting Started manual.
Download from Www.Somanuals.com. All Manuals Search And Download.
Coronis Fusion 4MP/6MP 37
Exclude I Guard zone
Scaling
When switched on, the video resolution of
the preferred input is set to a value with a
reduced number of lines. The image will be
displayed below the I-Guard patch.
When switched on, the display will enlarge
(scale up) low resolution video signals if
possible.
The DPMS system will power down the display when the
connected computer is left idle for a certain time.
Barco recommends to switch DPMS on to prevent image burn-
in (image retention) on the LCD panel..
Information menu
Product
The display type
Serial No
SW Number
SW Version
Indicates the display serial number
The order no. of the internal software
Displays the current internal software
version
Display Runtime
Indicates the total time the display has
been operating, including the time in
stand-by
Backlight Runtime
Indicates the total time the display has
been operating, excluding the time in
stand-by
Download from Www.Somanuals.com. All Manuals Search And Download.
38 Coronis Fusion 4MP/6MP
Changing the viewing mode
ABOUT VIEWING MODES
The display contains two viewing modes. Diagnostic viewing mode
provides the full calibrated luminance. This mode is intended for using
the display in diagnostic applications.
In Text mode, the luminance is approximately half of the luminance in
Diagnostic mode. This is intended for using the display with office
applications such as word processing.
SWITCHING VIEWING MODES BY MEANS OF THE KEYS
While the OSD is not on the screen, touch the Up and Down touch
keys at the same time. As a result the display switches to another
viewing mode. The viewing mode is indicated by a message that
appears for a few seconds after switching the viewing mode.
SWITCHING VIEWING MODES BY MEANS OF THE OSD MENUS
You can also switch viewing modes in the Luminance and color menu.
Enabling/disabling User Controls
TO DISABLE THE USER CONTROLS:
1 Open the OSD main menu.
2 Open the Settings menu.
3 Select User Controls.
4 Select Off.
5 Return to the Settings menu.
6 Exit the OSD
Now the user cannot display the OSD again until the user controls
keycode is entered (see below).
Download from Www.Somanuals.com. All Manuals Search And Download.
Coronis Fusion 4MP/6MP 39
TO ENABLE THE USER CONTROLS AGAIN USING THE TOUCH KEYS:
1 To enter the user controls keycode, the OSD must not be visible.
2 Touch any of the soft touch keys to switch on the front
illumination.
3 While the illumination is on, touch the following keys in successive
order:
Enter , Down , Up , Down , Enter
.
As a result, the User Controls setting in the Settings menu is switched
on again.
TO ENABLE THE USER CONTROLS AGAIN USING THE CONTROL WHEEL:
1 The OSD must not be visible.
2 Press and hold the control wheel for at least 5 seconds.
3 Rotate the control wheel 1 step clockwise.
4 Press the control wheel 2 times.
5 Rotate the control wheel 1 step counterclockwise. The on-screen
display appears.
Steps 3 to 5 must be performed in maximum 3 seconds.
As a result, the User Controls setting in the Settings menu is switched
on again.
Download from Www.Somanuals.com. All Manuals Search And Download.
40 Coronis Fusion 4MP/6MP
Switching Power LED on/off
TO SWITCH THE POWER LED ON/OFF:
1 Open the OSD main menu.
2 Open the Settings menu.
3 Select Power LED.
4 Select On or Off.
5 Return to the Settings menu.
When the display is switched in stand-by, the power LED will
turn orange, even if the power LED setting is switched off.
When the power LED setting is switched off, the LED will still
turn on (green) when the display is on but receives no video
signal.
Switching DPMS on/off
TO SWITCH DPMS ON/OFF:
1 Open the main menu.
2 Open the Settings menu.
3 Select DPMS.
4 Select On or Off.
5 Return to the Settings menu.
Download from Www.Somanuals.com. All Manuals Search And Download.
Coronis Fusion 4MP/6MP 41
Changing Display Functions
Concepts
• The Display Function menu lets you select a display function (DF).
• When you select the DICOM DF you can define the DICOM DF to be
recalculated taking the ambient light into account. This can be a
continuous recalculation (“Continuous DICOM ALC” on) or single
recalculation (“Continuous DICOM ALC” off).
• When you choose for a single recalculation, the recalculation can
be based on an illuminance of 0 lux (“Dark Room”) a preset value
(“Preset”) or a single, real-time measurement (“Measured ALC”).
Both Preset and Measured ALC correction values are determined by
the selected reading room.
• In the case of “Preset” or “Measured ALC” you must select a
reading room that corresponds to the type of room the display is
installed in. E.g. if the display is installed in a room where CT scans
are observed, the “CT/MR/NM” reading room is the preferred
selection.
• The reading room is defined by 3 parameters:
• Maximum Ambient Light defines the maximum light
allowed in this type of room. This value can be adjusted
within certain limits determined by the selected reading
room.
• Preset Correction Value is the predefined correction value
for this reading room. This value can be adjusted within
certain limits determined by the selected reading room.
• Measured ALC Value is the averaged measured ambient
light for this reading room. It can be changed by executing
the “Measure ALC Value” function.
• The reading room parameters are pre-defined in the display
according to the AAPM (American Association of Physicists in
Medicine). However they can be changed within certain limits.
Download from Www.Somanuals.com. All Manuals Search And Download.
42 Coronis Fusion 4MP/6MP
Display function selection
TO SELECT A DISPLAY FUNCTION (DF):
1 Open the main menu.
2 Open the Display Function menu.
3 Select Display Function.
4 Select the desired DF.
5 Return to the Display Function menu.
You can select the following DFs:
DICOM
Select a DICOM display function for most
medical viewing applications.
The DICOM function results in more visible
grayscales in the images.
Dynamic Gamma 2.2
Dynamic Gamma 1.8
These are gamma functions that are shifted
to take into account the non-zero
luminance of an LCD panel when driven
with a “black” signal.
They are especially useful in CT applications
to improve the perception of low
Hounsfield values.
Native
Test
If you select Native, the native panel
behavior will not be corrected.
This is identical to the Native DF.
Gamma 2.2
Gamma 1.8
Select one of these display functions in
case the display is to replace a CRT display
with a gamma of 1.8 or 2.2 respectively.
User
This is a user-definable DF.
Download from Www.Somanuals.com. All Manuals Search And Download.
Coronis Fusion 4MP/6MP 43
ALC & DICOM Options
Continuous DICOM ALC
INTRODUCTION
When you select the DICOM DF, the display offers the possibility to take
the ambient light into account for calculating the DICOM display function.
This function is called Continuous DICOM ALC (ALC is short for Ambient
Light Compensation).
When Continuous DICOM ALC is switched on, the display will
continuously adapt the DICOM DF based on the average measured
ambient light.
When it is switched off, the DICOM DF will be recalculated only when
another DICOM Offset is selected.
TO SWITCH CONTINUOUS DICOM ALC ON/OFF:
1 Select the DICOM display function. See To select a display
function (DF):, on page 43
2 In the Display Function menu, select ALC & DICOM Options.
3 In the ALC & DICOM Options submenu, select Cont. DICOM ALC.
4 Select On or Off.
5 Return to the ALC & DICOM Options menu.
The ambient light is measured by the ALC sensor at the front.
DICOM offset
INTRODUCTION
When Continuous DICOM ALC is switched off, the DICOM DF will be
recalculated only when another DICOM Offset is selected. The DICOM DF
will be recalculated based on the new offset setting.
Download from Www.Somanuals.com. All Manuals Search And Download.
44 Coronis Fusion 4MP/6MP
TO SELECT A DICOM OFFSET:
1 Select the DICOM display function. See To select a display
function (DF):, on page 43.
3 In the ALC & DICOM Options menu, select DICOM Offset.
4 Select an offset.
5 Return to the ALC & DICOM Options menu.
You can select the following offsets:
Dark room
The DICOM DF will be recalculated taking
an ambient light value of 0 lux into
account.
This means that the measured ambient
light does not influence the DICOM DF.
Preset
The DICOM DF will be recalculated taking a
preset ambient light value into account.
This value is determined by the selected
Reading Room and can be adjusted
manually within certain limits.
Measured ALC
The DICOM DF will be recalculated taking
the averaged measured ambient light
value into account that is determined by
the selected Reading Room. This value can
be changed by executing the “Measure ALC
Value” function in the Reading Room
Definition submenu.
Reading room selection
This function is available only when the DICOM display function is
selected and Continuous DICOM ALC is off.
Download from Www.Somanuals.com. All Manuals Search And Download.
Coronis Fusion 4MP/6MP 45
TO SELECT A READING ROOM SETTING:
1 Select the DICOM display function. See To select a display
function (DF):, on page 43.
3 In the ALC & DICOM Options menu, select Reading Room.
4 Select the reading room that corresponds to the room type the
display is installed in.
5 Return to the menu.
You can select the following reading room types:
DRR CR/DR/MAMMO
DRR CT/MR/NM
Corresponds to light conditions in
diagnostic reading rooms for computed
radiology, digital radiology or
mammography. This setting has the lowest
maximum ambient light.
Corresponds to light conditions in
diagnostic reading rooms for computed
tomography, magnetic resonance or
nuclear medicine scans.
Staff Office
Corresponds to light conditions in office
rooms.
Clinical Viewing
Corresponds to light conditions in
diagnostic reading rooms for clinical
viewing.
Emergency Room
Operating Room
Corresponds to light conditions in
emergency rooms.
Corresponds to light conditions in operating
rooms. This setting has the highest
maximum ambient light.
Download from Www.Somanuals.com. All Manuals Search And Download.
46 Coronis Fusion 4MP/6MP
Reading room definition
This function is available only when the DICOM display function is
selected and Continuous DICOM ALC is off.
TO MODIFY A READING ROOM DEFINITION:
1 Select the DICOM display function. See To select a display
function (DF):, on page 43.
3 In the ALC & DICOM Options menu, select Reading Room Def.
4 Select the reading room definition you wish to modify. Thereto,
first select Reading Room. Select the appropriate reading room.
5 To modify the maximum ambient light for this reading room,
select Max. Ambient Light and change the value.
You can change this value within certain limits determined by the
selected reading room.
6 To modify the Preset correction value for this reading room, select
Preset Corr. Value and change the value.
You can change this value within certain limits determined by the
selected reading room.
7 To modify the Measured ALC Value for this reading room, select
Measure ALC Value and execute the function.
If the “Averaged Ambient Light” at that moment is lower than the
defined “Maximum Ambient Light”, the averaged ambient light
value is copied as Measured ALC value.
If the “Averaged Ambient Light” at that moment is higher than the
defined “Maximum Ambient Light”, the display will prompt a
warning for the user to take measures to adapt the ambient light
in the room to the selected reading room. Proceed as follows:
• Use Enter
or click the control wheel. As a result the
Measured ALC value will be limited to the maximum ambient
Download from Www.Somanuals.com. All Manuals Search And Download.
Coronis Fusion 4MP/6MP 47
light value. Use Enter
store the value.
or click the control wheel again to
• However, if you wish to adapt the ambient light conditions to
the definition of the selected reading room, dim the light in
the room so that the measured ambient light becomes lower
that the maximum ambient light.
• Then use Enter
or click the control wheel again to execute
the Measure ALC Value function once more. As a result the
actually measured ambient light value will be copied as
Measured ALC Value. Finally use Enter
wheel again to store the value.
or click the control
Download from Www.Somanuals.com. All Manuals Search And Download.
48 Coronis Fusion 4MP/6MP
Cleaning your display
Front glass
Clean the front glass using a sponge, cleaning cloth or soft tissue, lightly
Before cleaning the front glass, unplug the power cable from
the mains to prevent the keys from being activated
inadvertently by sweeping over them.
moistened with one of the following tested products:
• Flux
• Pril
• Windex Glass Plus
• Bohle glass cleaner
• Mr. Proper
• Ajax glass cleaner
• Sidolin glass cleaner
• 25% ethanol solution (EtOH)
In case none of the above cleaning products is available, use plain water.
Take care not to damage or scratch the front glass. Be careful
with rings or other jewelry and do not apply excessive pressure
on the front glass.
Do not apply or spray liquid directly to the front glass as excess
liquid may cause damage to internal electronics. Instead, apply
the liquid to the cleaning cloth.
Download from Www.Somanuals.com. All Manuals Search And Download.
Coronis Fusion 4MP/6MP 49
Do not use following products:
• Alcohol/solvents at higher concentration > 5%
• Strong alkalis lye, strong solvents
• Acid
• Detergents with fluoride
• Detergents with ammonia
• Detergents with abrasives
• Steel wool
• Sponge with abrasives
• Steel blades
• Cloth with steel thread
Cabinet
1 Clean the cabinet using a soft cotton cloth, lightly moistened with
a recognized cleaning product for medical equipment.
The cabinet has been tested for resistance to the following
products:
• Cidex, Betadine
• Alcohol (Isopropyl and Ethyl)
• Ammonia-based cleaners (Windex)
• Aquasonic Gel
In case none of the above cleaning products is available, use plain
water.
Do not apply or spray liquid directly to the cabinet as
excess liquid may cause damage to internal electronics.
Instead, apply the liquid to the cleaning cloth.
2 Repeat with water only.
3 Wipe dry with a dry cloth.
Download from Www.Somanuals.com. All Manuals Search And Download.
50 Coronis Fusion 4MP/6MP
Important information
Safety information
GENERAL RECOMMENDATIONS
Read the safety and operating instructions before operating the display.
Retain safety and operating instructions for future reference.
Adhere to all warnings on the display and in the operating instructions
manual.
Follow all instructions for operation and use.
ELECTRICAL SHOCK
TYPE OF PROTECTION (ELECTRICAL):
Display with external power supply: Class I equipment
DEGREE OF SAFETY (FLAMMABLE ANESTHETIC MIXTURE):
Equipment not suitable for use in the presence of a flammable
anesthetic mixture with air or with oxygen or nitrous oxide.
NON-PATIENT CARE EQUIPMENT
Equipment primarily for use in a health care facility that is intended for
use where contact with a patient is unlikely.
Download from Www.Somanuals.com. All Manuals Search And Download.
Coronis Fusion 4MP/6MP 51
POWER CONNECTION - DISPLAY WITH EXTERNAL 24 VDC POWER
SUPPLY
• Power requirements: The display must be powered using the
delivered medical approved 24 VDC SELV power supply.
• The medical approved DC power supply must be powered by the
AC mains voltage (protective earth terminal).
TRANSIENT OVER-VOLTAGE
If the equipment is not used for a long time, disconnect it from the AC
inlet to avoid damage by transient over-voltage.
To fully disengage the power to the unit, please disconnect the power
cord from the AC inlet.
POWER CORDS:
• Utilize a UL-listed detachable power cord, 3-wire, type SJ or
equivalent, 18 AWG min., rated 300 V min., provided with a
hospital-grade type plug 5-15P configuration for 120V application,
or 6-15P for 240V application.
• Do not overload wall outlets and extension cords as this may result
in fire or electric shock.
• Mains lead protection (U.S.: Power cord): Power cords should be
routed so that they are not likely to be walked upon or pinched by
items placed upon or against them, paying particular attention to
cords at plugs and receptacles.
WATER AND MOISTURE
Never expose the display to rain or moisture.
Never use the display near water - e.g. near a bathtub, washbasin,
swimming pool, kitchen sink, laundry tub or in a wet basement.
VENTILATION
Do not cover or block the ventilation openings in the cover of the set.
When installing the display in a cupboard or another closed location,
Download from Www.Somanuals.com. All Manuals Search And Download.
52 Coronis Fusion 4MP/6MP
heed the necessary space between the set and the sides of the
cupboard.
INSTALLATION
Place the display on a flat, solid and stable surface that can support the
weight of at least 3 displays. If you use an unstable cart or stand, the
display may fall, causing serious injury to a child or adult, and serious
damage to the equipment.
THIS APPARATUS CONFORMS TO:
CE0120 (MDD 93/42/EEC; A1:2007/47/EC class IIb product), CE - 2004/
108/EC, CE - 93/42/EEC; A1:2007/47/EC class II b, IEC 60950-1:2005 +
A1:2009 (2ND EDITION), IEC 60601-1 2ND ED:1988 + A1:1991 + A2:1995,
UL 60601-1 1ST EDITION, CAN/CSA-C22.2 NO. 601.1-M90:2005, IEC
60601-1:2005 + A1:2012, ANSI/AAMI ES 60601-1:2005 + C1:2009 +
A1:2012, CAN/CSAC22.2 No. 60601-1-08:2008, DEMKO - EN 60601-
1:2006, EN 60601-1-2:2007, CCC - GB9254-2008 + GB4943-2001 +
GB17625.1-2003, KC, VCCI, FCC class B, ICES-001 Level B, FDA 510K,
RoHS.
NATIONAL SCANDINAVIAN DEVIATIONS FOR CL. 1.7.2:
Finland: "Laite on liitettävä suojamaadoituskoskettimilla varustettuun
pistorasiaan"
Norway: "Apparatet må tilkoples jordet stikkontakt"
Sweden: "Apparaten skall anslutas till jordat uttag"
Download from Www.Somanuals.com. All Manuals Search And Download.
Coronis Fusion 4MP/6MP 53
Environmental information
DISPOSAL INFORMATION (WASTE ELECTRICAL AND ELECTRONIC
EQUIPMENT)
This symbol on the product indicates that, under European
Directive 2002/96/EC governing waste from electrical and
electronic equipment, this product must not be disposed of
with other municipal waste. Please dispose of your waste
equipment by handing it over to a designated collection point
for the recycling of waste electrical and electronic equipment. To prevent
possible harm to the environment or human health from uncontrolled
waste disposal, please separate these items from other types of waste
and recycle them responsibly to promote the sustainable reuse of
material resources.
For more information about recycling of this product, please contact your
local city office, your municipal waste disposal service or the shop where
you purchased the product.
MERCURY NOTICE
This Barco product consists of materials that may contain mercury, which
must be recycled or disposed of in accordance with local, state, or
country laws:
• Within this system, the backlight lamps in the monitor display
contain mercury
TURKEY ROHS COMPLIANCE
Republic of Turkey: In conformity with the EEE Regulation
Türkiye Cumhuriyeti: EEE Yönetmeliğine Uygundur
Download from Www.Somanuals.com. All Manuals Search And Download.
54 Coronis Fusion 4MP/6MP
中国大陆 ROHS (CHINESE MAINLAND ROHS)
根据中国大陆 《电子信息产品污染控制管理办法》(也称为中国大
陆 RoHS),以下部分列出了 Barco 产品中可能包含的有毒和 / 或有
害物质的名称和含量。中国大陆 RoHS 指令包含在中国信息产业部
MCV 标准:“ 电子信息产品中有毒物质的限量要求 ” 中。
According to the "China Administration on Control of Pollution Caused by
Electronic Information Products" (Also called RoHS of Chinese Mainland),
the table below lists the names and contents of toxic and/or hazardous
substances that Barco's product may contain. The RoHS of Chinese
Mainland is included in the MCV standard of the Ministry of Information
Industry of China, in the section "Limit Requirements of toxic substances
in Electronic Information Products".
零件项目 ( 名称 )
有毒有害物质或元素
Component Name
Hazardous Substances or Elements
铅
汞
镉
六价铬
多溴联苯
多溴二苯醚
(Pb)
(Hg)
(Cd)
(Cr6+)
(PBB)
(PBDE)
印制电路配件
Printed Circuit
X
O
O
O
O
O
Assemblies
液晶面板
LCD panel
X
X
X
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
外接电 ( 线 ) 缆
External cables
內部线路
Internal wiring
O
O
O
O
O
金属外壳
Metal enclosure
塑胶外壳
Plastic enclosure
散热片 ( 器 )
Heatsinks
风扇
Fan
Download from Www.Somanuals.com. All Manuals Search And Download.
Coronis Fusion 4MP/6MP 55
电源供应器
Power supply unit
X
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
文件说明书
Paper manuals
光盘说明书
CD manual
O: 表示该有毒有害物质在该部件所有均质材料中的含量均在 SJ/T 11363-2006 标准规定的限
量要求以下 .
O: Indicates that this toxic or hazardous substance contained in all of the homogeneous
materials for this part is below the limit requirement in SJ/T11363-2006.
X: 表示该有毒有害物质至少在该部件的某一均质材料中的含量超出 SJ/T 11363-2006 标准规
定的限量要求 .
X: Indicates that this toxic or hazardous substance contained in at least one of the
homogeneous materials used for this part is above the limit requirement in SJ/T11363
2006.
在中国大陆销售的相应电子信息产品 (EIP)都必须遵照中国大陆
《电子信息产品污染控制标识要求》标准贴上环保使用期限
(EFUP)标签。Barco 产品所采用的 EFUP 标签 (请参阅实例,徽
标内部的编号使用于制定产品)基于中国大陆的 《电子信息产品
环保使用期限通则》标准。
All Electronic Information Products (EIP) that are sold within Chinese
Mainland must comply with the "Electronic Information Products
Pollution Control Labeling Standard" of Chinese Mainland, marked with
the Environmental Friendly Use Period (EFUP) logo. The number inside
the EFUP logo that Barco uses (please refer to the photo) is based on the
"Standard of Electronic Information Products Environmental Friendly Use
Period" of Chinese Mainland.
10
Download from Www.Somanuals.com. All Manuals Search And Download.
56 Coronis Fusion 4MP/6MP
Regulatory compliance information
INDICATIONS FOR USE: CORONIS FUSION 4MP
The Coronis Fusion 4MP (MDCC-4130) display is a display device capable
of displaying two 2MP display images electronically fused together on
the one panel. It is intended to be used in displaying and viewing digital
images for review and analysis by trained medical practitioners. It
provides a screen resolution of 2560x 1600 pixels at a pixel pitch of
0.2505mm. This provides the equivalent resolution of 2 x 2MP displays.
The Coronis Fusion 4MP (MDCC-4130) display is not intended for primary
diagnostic review of Mammography images.
Caution (USA): Federal law restricts this device to sale by or on the order
of a physician or a practitioner trained on its use.
INDICATIONS FOR USE: CORONIS FUSION 6MP
The Coronis Fusion 6MP DL (MDCC-6130) display is a display device
capable of displaying two 3MP display images electronically fused
together on the one panel. It is intended to be used in displaying and
viewing digital images for review and analysis by trained medical
practitioners. It provides a screen resolution of 3280 x 2048 pixels at a
pixel pitch of 0.1995mm. This provides the equivalent resolution of 2 x
3MP displays.
The Coronis Fusion 6MP DL (MDCC-6130) display is not intended for
primary diagnostic review of Mammography images.
Caution (USA): Federal law restricts this device to sale by or on the order
of a physician or a practitioner trained on its use.
FCC CLASS B
This device complies with Part 15 of the FCC Rules. Operation is subject
to the following two conditions: (1) this device may not cause harmful
interference, and (2) this device must accept any interference received,
including interference that may cause undesired operation.
This equipment has been tested and found to comply with the limits for
a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits
Download from Www.Somanuals.com. All Manuals Search And Download.
Coronis Fusion 4MP/6MP 57
are designed to provide reasonable protection against harmful
interference in a residential installation. This equipment generates, uses
and can radiate radio frequency energy and, if not installed and used in
accordance with the instructions, may cause harmful interference to
radio communications. However, there is no guarantee that interference
will not occur in a particular installation. If this equipment does cause
harmful interference to radio or television reception, which can be
determined by turning the equipment off and on, the user is encouraged
to try to correct the interference by one or more of the following
measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from
that to which the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
CANADIAN NOTICE
This ISM device complies with Canadian ICES-001.
Cet appareil ISM est conforme à la norme NMB-001 du Canada.
Download from Www.Somanuals.com. All Manuals Search And Download.
58 Coronis Fusion 4MP/6MP
EMC notice
No specific requirement on the use of external cables or other
accessories except power supply.
With the installation of the device, use only the delivered power supply
or a spare part provided by the legal manufacturer. Using another can
result in a decrease of the immunity level of the device.
ELECTROMAGNETIC EMISSIONS
The CORONIS FUSION 4MP/6MP display is intended for use in the
electromagnetic environment specified below.
The customer or the user of the CORONIS FUSION 4MP/6MP display should
assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
Compliance
Electromagnetic environment – guidance
Group 1
The CORONIS FUSION 4MP/6MP display uses
RF energy only for its internal function.
Therefore, its RF emissions are very low and
are not likely to cause any interference in
nearby electronic equipment
RF emissions
CISPR 11
Class B
Class D
The CORONIS FUSION 4MP/6MP display is
suitable for use in all establishments,
including domestic establishments and
those directly connected to the public low-
voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
Complies
IEC 61000-3-3
This CORONIS FUSION 4MP/6MP display complies with appropriate medical
EMC standards on emissions to, and interference from surrounding
equipment. Operation is subject to the following two conditions: (1) this
device may not cause harmful interference, and (2) this device must
Download from Www.Somanuals.com. All Manuals Search And Download.
Coronis Fusion 4MP/6MP 59
accept any interference received, including interference that may cause
undesired operation.
Interference can be determined by turning the equipment off and on.
If this equipment does cause harmful interference to, or suffer from
harmful interference of, surrounding equipment, the user is encouraged
to try to correct the interference by one or more of the following
measures:
• Reorient or relocate the receiving antenna or equipment.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from
that to which the receiver is connected.
• Consult the dealer or an experienced technician for help
Download from Www.Somanuals.com. All Manuals Search And Download.
60 Coronis Fusion 4MP/6MP
ELECTROMAGNETIC IMMUNITY
The CORONIS FUSION 4MP/6MP display is intended for use in the
electromagnetic environment specified below. The customer or the user
of the CORONIS FUSION 4MP/6MP display should assure that it is used in
such an environment.
Immunity test
IEC 60601
Test levels
Compliance
level
Electromagnetic
environment –
guidance
Electrostatic
discharge (ESD)
± 6kV contact
± 8kV air
± 6kV contact
± 8kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%
IEC 61000-4-2
Electrical fast
transient/burst
± 2kV for power
supply lines
± 2kV for power
supply lines
Mains power quality
should be that of a
typical commercial or
hospital environment
IEC 61000-4-4
± 1kV for input/
output lines
± 1kV for input/
output lines
Surge
± 1 kV line(s) to
line(s)
± 1 kV line(s) to
line(s)
Mains power quality
should be that of a
typical commercial or
hospital environment
IEC61000-4-5
± 2 kV line(s) to
earth
± 2 kV line(s) to
earth
Voltage dips,
short
< 5% UT (> 95%
dip in UT) for 0.5
cycle
< 5% UT (> 95%
dip in UT) for 0.5
cycle
Mains power quality
should by that of a
typical commercial or
hospital environment. If
interruptions
and voltage
variations on
power supply
input lines
40% UT (60%
dip in UT) for 5
cycles
40% UT (60% dip
in UT) for 5 cycles
the user of the CORONIS
FUSION 4MP/6MP display
requires continued
operation during power
mains interruptions, it is
recommended that the
70% UT (30% dip
in UT) for 25
cycles
IEC 61000-4-11
70% UT (30%
dip in UT) for 25
cycles
CORONIS FUSION 4MP/6MP
< 5% UT (>95%
dip in UT) for 5s
< 5% UT (>95%
dip in UT) for 5s
display be powered from
an uninterruptible power
supply or a battery.
UT is the a.c. mains voltage prior to application of the test level
Download from Www.Somanuals.com. All Manuals Search And Download.
Coronis Fusion 4MP/6MP 61
Immunity
test
IEC 60601
test level
Level
Electromagnetic environment –
guidance
Power
3 A/m
Not
applic
able
Power frequency magnetic fields should be
at levels characteristic of a typical location
in a typical commercial or hospital
environment.
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Conducted RF
IEC 61000-4-6
3 Vrms
3 V
Portable and mobile RF communications
equipment should be used no closer to any
part of the CORONIS FUSION 4MP/6MP
display, including cables, than the
150 kHz to
80 MHz
recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
3 V/
m
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to
2.5 GHz
Recommended separation distance
d 1.2
d 1.2
d 2.3
P
P
P
80 MHz to 800 MHz
800 MHz to 2.5 GHz
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey,a should be less than the
compliance level in each frequency range.b
Interference may occur in the vicinity of
equipment marked with symbol:
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the CORONIS FUSION 4MP/
Download from Www.Somanuals.com. All Manuals Search And Download.
62 Coronis Fusion 4MP/6MP
6MP display is used exceeds the applicable RF compliance level above, the CORONIS FUSION
4MP/6MP display should be observed to verify normal operation. If abnormal performance
is observed, additional measures may be necessary, such as re-orienting or relocating the
CORONIS FUSION 4MP/6MP display.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
At 80 MHz and 800 MHz, the higher frequency range applies.
These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Download from Www.Somanuals.com. All Manuals Search And Download.
Coronis Fusion 4MP/6MP 63
RECOMMENDED SEPARATION DISTANCE
The CORONIS FUSION 4MP/6MP display is intended for use in an
electromagnetic environment in which radiated RF disturbances are
controlled. The customer of the user of the CORONIS FUSION 4MP/6MP
display can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications
equipment (transmitters) and the CORONIS FUSION 4MP/6MP display as
recommended below,
according to the maximum output power of the communications
equipment.
Rated
Separation distance according to frequency of transmitter
maximum
output power
of transmitter
150kHz to 80MHz
80MHz to 800MHz
800MHz to 2.5GHz
W
d 1.2
P
d 1.2
P
d 2.3 P
0.01
0.1
1
0.12
0.38
1.2
0.12
0.38
1.2
0.23
0.73
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter. Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
At 80MHz and 800MHz, the separation distance for the higher
frequency range applies.
These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and
reflection form structures, object and people.
Download from Www.Somanuals.com. All Manuals Search And Download.
64 Coronis Fusion 4MP/6MP
Explanation of symbols
On the display or power supply, you may find the following symbols
(nonrestrictive list):
Indicates compliance to the essential
requirements of the Directive 93/42/EEC
0120
Indicates compliance with Part 15 of the
FCC rules (Class A or Class B)
Indicates the display is approved according
to the UL regulations
or
Indicates the display is approved according
to the c-UL regulations
or
Indicates the display is approved according
to the DEMKO regulations
Indicates the display is approved according
S&E
to the CCC regulations
Indicates the display is approved according
to the VCCI regulations
Indicates the USB connectors on the display
Indicates the manufacturing date
Indicates the temperature limitations for
the display to operate within specs
Indicates the display serial no.
Download from Www.Somanuals.com. All Manuals Search And Download.
Coronis Fusion 4MP/6MP 65
Consult the operating instructions
Indicates this apparatus must not be
thrown in the trash but must be recycled,
according to the European WEEE (Waste
Electrical and Electronic Equipment)
directive
Legal disclaimer
DISCLAIMER NOTICE
Although every attempt has been made to achieve technical accuracy in
this document, we assume no responsibility for errors that may be
found. Our goal is to provide you with the most accurate and usable
documentation possible; if you discover errors, please let us know.
Barco software products are the property of Barco. They are distributed
under copyright by Barco N.V. or Barco, Inc., for use only under the
specific terms of a software license agreement between Barco N.V. or
Barco Inc. and the licensee. No other use, duplication, or disclosure of a
Barco software product, in any form, is authorized.
The specifications of Barco products are subject to change without
notice.
TRADEMARKS
All trademarks and registered trademarks are property of their
respective owners.
COPYRIGHT NOTICE
This document is copyrighted. All rights are reserved. Neither this
document, nor any part of it, may be reproduced or copied in any form
or by any means - graphical, electronic, or mechanical including
Download from Www.Somanuals.com. All Manuals Search And Download.
66 Coronis Fusion 4MP/6MP
photocopying, taping or information storage and retrieval systems -
without written permission of Barco.
© 2013 Barco N.V. All rights reserved.
PATENT INFORMATION
This product is covered under the following intellectual property rights:
US Patent RE43,707
US Patent 7,038,186
US Patent 7,166,829
US Patent 6,950,098
European Patent 1 274 066.
Download from Www.Somanuals.com. All Manuals Search And Download.
Coronis Fusion 4MP/6MP 67
Technical specifications
CORONIS FUSION 4MP
Product acronym
Screen technology
MDCC-4130
TFT AM Color LCD Dual Domain IPS-Pro
758.36 mm (29.9")
Active screen size
(diagonal)
Active screen size (H x V)
Aspect ratio (H:V)
Resolution
641.208 x 404.808 mm (25.25 x 15.94")
16:10
native 4MP (2560 x 1600)
configurable to 2 x 2MP+ (1280 x 1600)
configurable to 2 x 2MP (1200 x 1600)
Pixel pitch
0.2505 mm
Yes
Color imaging
Gray imaging
Color support
Viewing angle (H, V)
Yes
30 bit
170°
Uniform Luminance
Technology (ULT)
Yes
Per Pixel Uniformity (PPU)
No
Ambient Light
Yes
Compensation (ALC)
Backlight Output
Yes (1x)
Stabilization (BLOS)
I-Guard
Yes
Maximum luminance
1000 cd/m²
500 cd/m²
DICOM calibrated
luminance (ULT off)
Contrast ratio (typical)*
1100:1
Download from Www.Somanuals.com. All Manuals Search And Download.
68 Coronis Fusion 4MP/6MP
Response time (Tr + Tf)
Scanning frequency (H; V)
Housing color
20 ms
30-150 kHz; 15-80 Hz
Black/Silver
Video input signals
Video inout terminals
USB ports
DVI-D Dual Link (2x)
NA
1 upstream (endpoint), 3 downstream
USB standard
2.0
Power requirements
(nominal)
100-240V
Power consumption
(nominal)
135W
Power save mode
Power management
Dot clock
Yes
DVI-DMPM
280 MHz
OSD languages
English, French, German, Spanish, Italian
Dimensions with stand (W
x H x D)
Portrait: NA
Landscape: 731 x 580~676 x 265 mm
Dimensions w/o stand (W
x H x D)
Portrait: NA
Landscape: 731 x 485 x 141 mm
Dimensions packaged (W x
H x D)
869 x 764 x 400 mm
Net weight with stand
Net weight w/o stand
28.1 kg
21.6 kg
37.6 kg
Net weight packaged with
stand
Net weight packaged w/o
stand
NA
Height adjustment range
Tilt
96 mm
-5° / +25°
Download from Www.Somanuals.com. All Manuals Search And Download.
Coronis Fusion 4MP/6MP 69
Swivel
-45° / +45°
Pivot
NA
Mounting standard
Screen protection
Recommended modalities
Certifications
VESA (100 mm)
Protective, non-reflective glass cover
CT, MR, US, DR, CR, NM, Film
CE0120 (MDD 93/42/EEC; A1:2007/47/EC class IIb
product), CE - 2004/108/EC, CE - 93/42/EEC;
A1:2007/47/EC class II b, IEC 60950-1:2005 +
A1:2009 (2ND EDITION), IEC 60601-1 2ND ED:1988 +
A1:1991 + A2:1995, UL 60601-1 1ST EDITION, CAN/
CSA-C22.2 NO. 601.1-M90:2005, IEC 60601-1:2005 +
A1:2012, ANSI/AAMI ES 60601-1:2005 + C1:2009 +
A1:2012, CAN/CSAC22.2 No. 60601-1-08:2008,
DEMKO - EN 60601-1:2006, EN 60601-1-2:2007, CCC -
GB9254-2008 + GB4943-2001 + GB17625.1-2003, KC,
VCCI, FCC class B, ICES-001 Level B, FDA 510K, RoHS
Supplied accessories
Getting Started Guide
Quick-Installation Sheet
Video cables (2 x DVI Dual Link)
Main cables (UK, European (CEBEC/KEMA), USA (UL/
CSA; adaptor plug NEMA 5-15P), Chinese (CCC))
USB 2.0 cable
External power supply
Optional accessories
QA software
NA
MediCal QAWeb
Unit per pallet
NA
Pallet dimensions
Warranty
NA
5 years
Operating temperature
Storage temperature
Operating humidity
Storage humidity
Operating altitude
0°C to 35°C (15°C to 30°C within specs)
-20°C to 60°C
8% - 80% (non-condensing)
5% - 95% (non-condensing)
3000 m
* In dark reading room conditions (0 lux)
Download from Www.Somanuals.com. All Manuals Search And Download.
70 Coronis Fusion 4MP/6MP
CORONIS FUSION 6MP
Product acronym
Screen technology
MDCC-6130 DL
TFT AM Color LCD Dual Domain IPS-Pro
772 mm (30.4")
Active screen size
(diagonal)
Active screen size (H x V)
Aspect ratio (H:V)
Resolution
654 x 409 mm (25.8 x 16.1")
16:10
native 6MP (3280 x 2048)
configurable to 2 x 3MP+ (1640 X 2048)
configurable to 2 x 3MP (1536 X 2048)
Pixel pitch
0.1995 mm
Yes
Color imaging
Gray imaging
Color support
Viewing angle (H, V)
Yes
30 bit
178°
Uniform Luminance
Technology (ULT)
Yes
Per Pixel Uniformity (PPU)
No
Ambient Light
Yes
Compensation (ALC)
Backlight Output
Yes (2x)
Stabilization (BLOS)
I-Guard
Yes
Maximum luminance*
800 cd/m²
500 cd/m²
DICOM calibrated
luminance (ULT off)*
Contrast ratio (typical)*
Response time (Tr + Tf)
1000:1
18 ms
Download from Www.Somanuals.com. All Manuals Search And Download.
Coronis Fusion 4MP/6MP 71
Scanning frequency (H; V)
Housing color
30-150 kHz; 15-80 Hz
Black/Silver
Video input signals
Video inout terminals
USB ports
DVI-D Dual Link (2x)
NA
1 upstream (endpoint), 3 downstream
USB standard
2.0
Power requirements
(nominal)
100-240V
Power consumption
(nominal)
166W
Power save mode
Power management
Dot clock
Yes
DVI-DMPM
280 MHz
OSD languages
English, French, German, Spanish, Italian
Dimensions with stand (W
x H x D)
Portrait: NA
Landscape: 731 x 580~676 x 265 mm
Dimensions w/o stand (W
x H x D)
Portrait: NA
Landscape: 731 x 485 x 141 mm
Dimensions packaged (W x
H x D)
869 x 764 x 400 mm
Net weight with stand
Net weight w/o stand
26.9 kg
20.4 kg
36.1 kg
Net weight packaged with
stand
Net weight packaged w/o
stand
NA
Height adjustment range
Tilt
96 mm
-5° / +25°
Download from Www.Somanuals.com. All Manuals Search And Download.
72 Coronis Fusion 4MP/6MP
Swivel
-45° / +45°
Pivot
NA
Mounting standard
Screen protection
Recommended modalities
Certifications
VESA (100 mm)
Protective, non-reflective glass cover
CT, MR, US, DR, CR, NM, Film
CE0120 (MDD 93/42/EEC; A1:2007/47/EC class IIb
product), CE - 2004/108/EC, CE - 93/42/EEC;
A1:2007/47/EC class II b, IEC 60950-1:2005 +
A1:2009 (2ND EDITION), IEC 60601-1 2ND ED:1988 +
A1:1991 + A2:1995, UL 60601-1 1ST EDITION, CAN/
CSA-C22.2 NO. 601.1-M90:2005, IEC 60601-1:2005 +
A1:2012, ANSI/AAMI ES 60601-1:2005 + C1:2009 +
A1:2012, CAN/CSAC22.2 No. 60601-1-08:2008,
DEMKO - EN 60601-1:2006, EN 60601-1-2:2007, CCC -
GB9254-2008 + GB4943-2001 + GB17625.1-2003, KC,
VCCI, FCC class B, ICES-001 Level B, FDA 510K, RoHS
Supplied accessories
Getting Started Guide
Quick-Installation Sheet
Video cables (2 x DVI Dual Link)
Main cables (UK, European (CEBEC/KEMA), USA (UL/
CSA; adaptor plug NEMA 5-15P), Chinese (CCC))
USB 2.0 cable
External power supply
Optional accessories
QA software
NA
MediCal QAWeb
Unit per pallet
NA
Pallet dimensions
Warranty
NA
5 years
Operating temperature
Storage temperature
Operating humidity
Storage humidity
Operating altitude
0°C to 35°C (15°C to 30°C within specs)
-20°C to 60°C
8% - 70% (non-condensing)
5% - 95% (non-condensing)
3000 m
* In dark reading room conditions (0 lux)
Download from Www.Somanuals.com. All Manuals Search And Download.
Coronis Fusion 4MP/6MP 73
Download from Www.Somanuals.com. All Manuals Search And Download.
74 Coronis Fusion 4MP/6MP
Download from Www.Somanuals.com. All Manuals Search And Download.
Coronis Fusion 4MP/6MP 75
K5902015-08
February 2013
Barco n.v.
President Kennedypark 35
8500 Kortrijk
Belgium
0120
Download from Www.Somanuals.com. All Manuals Search And Download.
|