Barco Projector Accessories 4MP User Manual

Coronis Fusion 4MP/6MP  
User Guide  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Welcome!  
Thank you for choosing this CORONIS FUSION 4MP/6MP display!  
Coronis Fusion is an ingenious PACS display system designed to further  
enhance flexibility and productivity in diagnostic imaging.  
Featuring the industry’s first 30-inch color LCD that can be used as two  
seamless heads or one wide-screen display, Coronis Fusion offers you  
the freedom to organize your workspace just the way you want it. It  
allows you to read CT, MR, cath and echo cardiogram images, or any  
other combination, side by side on a single diagnostic screen.  
Coronis Fusion features the latest, breakthrough In-Plane Switching (IPS-  
Pro) LCD technology. This technology brings the most advanced LCD  
viewing characteristics on the market, setting new standards for  
brightness and contrast, even from a wide viewing angle.  
Use the instructions in this guide to install your CORONIS FUSION 4MP/6MP  
display and discover all these interesting features yourself!  
Following symbols may be used throughout this guide:  
Warning: Risk of  
injury to human  
beings  
Caution: Risk of  
damage to the  
product  
WARNING  
Important notice  
or remark  
Note  
Hint, tip  
Additional  
information  
Read all the safety information starting on page 51 before  
operating your CORONIS FUSION 4MP/6MP display.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Coronis Fusion 4MP/6MP  
3
 
Download from Www.Somanuals.com. All Manuals Search And Download.  
Coronis Fusion 4MP/6MP  
4
Table of contents  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Coronis Fusion 4MP/6MP  
5
 
Download from Www.Somanuals.com. All Manuals Search And Download.  
Coronis Fusion 4MP/6MP  
6
What's in the box  
Your CORONIS FUSION 4MP/6MP display comes with:  
this CORONIS FUSION 4MP/6MP display guide  
a system CD  
two DVI cables  
a USB cable  
a set of AC power cords  
an external power supply  
If you ordered a Barco display controller, it’s also in the box together  
with its accessories. A dedicated user guide is available on the system  
CD.  
Keep your original packaging. It is designed for this display and  
is the ideal protection during transport.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Coronis Fusion 4MP/6MP  
7
 
Parts, controls and connectors  
Front view  
1
2
3
4
5 6 7  
8
1
3
5
7
2
4
6
8
USB downstream connectors  
Right/Up key  
Left/Down key  
Enter key  
Stand-by key  
Ambient light sensor  
Control wheel  
Power status indicator light  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Coronis Fusion 4MP/6MP  
8
   
Rear view  
1
2
3
1
3
2
Connector compartment  
cover  
Neck cover  
Display stand cover  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Coronis Fusion 4MP/6MP  
9
 
CORONIS FUSION 4MP  
1
2
3
4
5
6
7
1
3
2
4
+24 VDC power input  
USB upstream connector  
USB downstream connector  
+5 VDC, 0.25 A power output  
for accessories  
5
7
6
+3.3 VDC, 0.25A auxiliary  
power output  
DVI 1 video input  
DVI 2 video input  
Download from Www.Somanuals.com. All Manuals Search And Download.  
10 Coronis Fusion 4MP/6MP  
CORONIS FUSION 6MP  
1
2
3
4
5
6
7
1
3
2
4
+24 VDC power input  
USB upstream connector  
USB downstream connector  
+5 VDC, 0.25 A power output  
for accessories  
5
7
6
+3.3 VDC, 0.25A auxiliary  
power output  
DVI 1 video input  
DVI 2 video input  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Coronis Fusion 4MP/6MP 11  
Display installation  
Prior to installing your CORONIS FUSION 4MP/6MP display and  
connecting all necessary cables, make sure to have a suitable  
display controller physically installed in your computer. If you are  
using a Barco display controller, please consult the dedicated user  
guide available on the system CD.  
Removing the covers  
CONNECTOR COMPARTMENT COVER  
The connector compartment cover should be removed to get access to  
the connectors. To remove the cover, proceed as follows:  
1 Gently lift the clips at one of the handles of the connector  
compartment cover to release that side of the cover.  
2 Do the same at the other side of the cover.  
3 Remove the cover.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
12 Coronis Fusion 4MP/6MP  
   
DISPLAY STAND COVER  
To get access to the cable routing channel, the display stand cover  
should be removed. This can be done as follows:  
1 Gently press and hold the clips at the inside top of the cover.  
2 Slide the cover upwards while holding the clips pressed.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Coronis Fusion 4MP/6MP 13  
Unlocking the height mechanism  
In the factory, the height-positioning system in the display stand is  
locked with a red hook to prevent damage during transportation. You’ll  
have to remove this hook before adjusting your display height position.  
1 While holding the display panel pushed down, pull out the red  
hook in the display stand.  
To retain the hook for possible future transportation, insert the  
short, red end of the hook back into the stand of your display.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
14 Coronis Fusion 4MP/6MP  
 
Adjusting the display position  
Now that the height-positioning system of your display is unlocked, you  
can tilt, swivel, raise and lower the display as desired.  
Do not try to pivot your display when attached to the stand.  
Trying to do so could cause serious damage to your display and  
its stand.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Coronis Fusion 4MP/6MP 15  
 
Connecting the signal cables  
1 Connect one head of the display controller to the DVI 1 connector  
with one of the supplied DVI cables.  
2 Connect another head of the display controller to the DVI 2  
connector with one of the supplied DVI cables.  
3 If you want to make use of your display’s USB downstream  
connectors, connect a PC USB downstream connector to the  
display’s USB upstream connector by means of the supplied USB  
2.0 cable.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
16 Coronis Fusion 4MP/6MP  
 
Connecting the power cables  
1 Connect the supplied external DC power supply to the +24 VDC  
power input of your CORONIS FUSION 4MP/6MP display.  
2 Plug the other end of the external DC power supply into a  
grounded power outlet by means of the proper power cord  
delivered in the packaging.  
Routing the cables  
Route all connected cables trough the cable routing channel in the stand  
of your display.  
The cable straps at the inside of the connector compartment  
allow you to fix the cables for better shielding of the cables.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Coronis Fusion 4MP/6MP 17  
   
Re-attaching the covers  
1 Re-attach the display stand cover by sliding it downward. You’ll  
hear a “click” sound of the cover’s clips when the display stand  
cover is in position.  
Pay attention that all cables stay in the cable channel  
while replacing the cover.  
2 Re-attach the connector compartment cover by sliding the cover’s  
top in position and then push the cover’s bottom. You’ll hear a  
“click” sound of the cover’s clips when the connector compartment  
cover is in position.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
18 Coronis Fusion 4MP/6MP  
 
VESA-mount installation  
The panel, standard attached to a stand, is compatible with the VESA  
100 mm standard. Thus, it can be used with an arm that is approved by  
VESA. This chapter shows you how to release the panel from the stand  
and how to attach it to an arm. If you’re not using an arm, you can skip  
this chapter.  
1 Gently pull open both lower sides of the cover.  
2 Slide the cover upwards while holding the lower sides pulled open.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Coronis Fusion 4MP/6MP 19  
 
3 Lift the plastic frame that covers the fixation of the panel to the  
stand. Turn it for 45 degrees to uncover the fixation screws.  
4 Unscrew the four fixation screws fixing the panel to the stand.  
5 Attach the panel firmly to the arm using 4 screws M4 x 10 mm.  
You should mount the panel in landscape position.  
Portrait position is possible but not supported.  
Use an arm that can support a weight of at least 21.6 kg  
(47.62 lbs). Failure to do so could make the panel fall,  
causing serious injury to a child or adult, and serious  
damage to the equipment.  
Never move a display attached to an arm by pulling or  
pushing the display itself. Instead, make sure that the  
arm is equipped with a VESA approved handle and use  
this to move the display.  
Please refer to the instruction manual of the arm for more  
information and instructions.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
20 Coronis Fusion 4MP/6MP  
First time starting up  
You are now ready to start up your CORONIS FUSION 4MP/6MP display  
system for the first time.  
1 Switch on your CORONIS FUSION 4MP/6MP display as described in  
2 Turn on the computer connected to your display.  
If you have properly installed your display and display controller, the  
Windows start-up messages will appear once the boot procedure is  
finished.  
Your CORONIS FUSION 4MP/6MP display will be running in a basic  
video mode at a default refresh rate when first time starting  
up.  
If you are using a Barco display controller, please consult the  
dedicated user guide available on the system CD to install the  
drivers, software and documentation. When this is done, your  
display will automatically detect the connected video input  
signal(s) and apply the correct video mode and refresh rate.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Coronis Fusion 4MP/6MP 21  
 
Daily operation  
Recommendations for daily operation  
OPTIMIZE THE LIFETIME OF YOUR DISPLAY  
Enabling the Display Power Management System (DPMS) of your display  
will optimize its diagnostic lifetime by automatically switching off the  
backlight when the display is not used for a specified period of time. By  
default, DPMS is enabled on your display, but it also needs to be  
activated on your workstation. To do this, go to “Power Options  
Properties” in the “Control Panel”.  
Barco recommends setting DPMS activation after 20 minutes of  
non-usage.  
USE A SCREEN SAVER TO AVOID IMAGE RETENTION  
Prolonged operation of an LCD with the same content on the same  
screen area may result in a form of image retention.  
You can avoid or significantly reduce the occurrence of this phenomenon  
by using a screen saver. You can activate a screen saver in the “Display  
properties” window of your workstation.  
Barco recommends setting screen saver activation after 5  
minutes of non-usage. A good screen saver displays moving  
content.  
In case you are working with the same image or an application with  
static image elements for several hours continuously (so that the screen  
saver is not activated), change the image content regularly to avoid  
image retention of the static elements.  
UNDERSTAND PIXEL TECHNOLOGY  
LCD displays use technology based on pixels. As a normal tolerance in  
the manufacturing of the LCD, a limited number of these pixels may  
remain either dark or permanently lit, without affecting the diagnostic  
Download from Www.Somanuals.com. All Manuals Search And Download.  
22 Coronis Fusion 4MP/6MP  
   
performance of the product. To ensure optimal product quality, Barco  
applies strict selection criteria for its LCD panels.  
To learn more about LCD technology and missing pixels, consult  
medical.  
ENHANCE USER COMFORT  
Every Barco multi-head display system is color matched with the highest  
specification in the market.  
Barco recommends keeping color-matched displays together.  
Furthermore, it is important to use all displays of a multi-head  
configuration at the same rate to preserve color matching  
throughout the economic lifetime of the system.  
MAXIMIZE QUALITY ASSURANCE  
The ‘MediCal QAWeb’ system offers online service for high-grade Quality  
Assurance, providing maximum diagnostic confidence and uptime.  
Barco recommends to install MediCal QAWeb Agent and apply  
the default QAWeb policy at least. This policy includes  
calibration on regular intervals. Upgrading to MediCal QAWeb  
Server offers even more possibilities.  
Learn more and sign up for the free MediCal QAWeb Essential  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Coronis Fusion 4MP/6MP 23  
Stand-by switching  
The connected power supply also provides a switch that can be  
used to turn the power completely off. To use the display,  
please make sure to switch on this power supply. This can be  
done by pushing the on/off switch on the power supply into  
the "|" position.  
USING THE TOUCH KEYS:  
1 While the OSD is not on the screen, touch any touch key. As a  
result, the control panel illumination is switched on for a few  
seconds.  
2 While the contol panel illumination is on, touch the stand-by  
touch key. A message appears on the screen.  
3 Touch the stand-by touch key once more to confirm.  
4 To switch the display on while it is in stand-by, touch any touch  
key. Next, while the contol panel illumination is on, touch the  
stand-by touch key.  
USING THE CONTROL WHEEL:  
1 While the OSD is not on the screen, click the control wheel shortly.  
A message appears on the screen.  
2 Click the control wheel once more to confirm.  
3 To switch the display on while it is in stand-by, click the control  
wheel shortly.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
24 Coronis Fusion 4MP/6MP  
 
Bringing up the OSD menu  
USING THE TOUCH KEYS:  
1 While the OSD is not on the screen, touch any touch key. As a  
result, the control panel illumination is switched on for a few  
seconds.  
2 While the contol panel illumination is on, touch the Enter  
touch  
key.  
USING THE CONTROL WHEEL:  
While the OSD is not on the screen, rotate the control wheel.  
Navigating through the OSD menus  
USING THE TOUCH KEYS:  
1 Use the Up and Down keys to move through the menus.  
2 To go into a submenu, use the Up or Down key to highlight  
the submenu. Next, touch the Enter  
submenu.  
key to go into the  
3 When you select Exit, the OSD exits the submenu and jumps to the  
previous (higher-level) menu. When you select Exit in the main  
menu, the OSD is hidden.  
USING THE CONTROL WHEEL:  
1 Rotate the control wheel to move up and down through the  
menus.  
2 To go into a submenu, rotate the control wheel to highlight the  
submenu. Next, click the control wheel to go into the submenu.  
3 When you click on Exit, the OSD exits the submenu and jumps to  
the previous (higher-level) menu. When you click on Exit in the  
main menu, the OSD is hidden.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Coronis Fusion 4MP/6MP 25  
   
Making selections in the OSD menus  
USING THE TOUCH KEYS:  
1 Navigate through the menus and highlight the parameter you wish  
to select.  
2 Touch the Enter  
key to make a selection or toggle between  
several settings.  
USING THE CONTROL WHEEL:  
1 Navigate through the menus and highlight the parameter you wish  
to select.  
2 Click the control wheel to make a selection or to toggle between  
several settings.  
Changing values in the OSD menus  
USING THE TOUCH KEYS:  
1 Navigate through the menus and highlight the parameter you wish  
to adjust.  
2 Touch the Enter  
key to go into edit mode.  
3 Use the Up or Down key to change the value.  
4 Touch the Enter  
key to enter the new value and return to the  
menu.  
USING THE CONTROL WHEEL:  
1 Navigate through the menus and highlight the parameter you wish  
to adjust.  
2 Click the control wheel to go into edit mode.  
3 Rotate the wheel to change the value.  
4 Click the wheel to enter the new value and return to the menu.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
26 Coronis Fusion 4MP/6MP  
   
Advanced operation  
Info messages  
The following messages may appear automatically on the screen:  
INPUT SELECTION  
The Input selection menu appears in case 2 video signals are connected  
and the resolution of one or both video signals is too high to display  
them both.  
In that case the input selection menu allows you to select one of both  
video signals.  
NO SIGNAL SELECTED  
This message appears in case the display finds no proper video signal at  
its inputs.  
RESOLUTION TOO HIGH  
This message appears in case the resolution of the selected video signal  
is too high to display.  
TEXT MODE / DIAGNOSTIC MODE  
These messages appear for a few seconds directly after switching to Text  
mode or Diagnostic mode.  
In text mode, the display luminance is approximately half of the  
luminance in diagnostic mode. Text mode is intended for using the  
display with office applications such as word processing. You can switch  
between both modes by touching the up and down keys at the  
same time.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Coronis Fusion 4MP/6MP 27  
   
LCM TEMPERATURE TOO HIGH  
This is a serious error message. You should power off the  
display and call for service.  
The warning means the LCD panel temperature is too high. As a result  
the display switches to a safe operating mode (with decreased backlight  
value). The message will not disappear until the display is powered off.  
LCM TEMPERATURE SENSOR OFF LINE  
This is a serious error message. You should power off the  
display and call for service.  
The warning means the LCD panel temperature sensor has lost  
communication. As a result the display switches to a safe operating  
mode (with decreased backlight value). The message will not disappear  
until the display is powered off.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
28 Coronis Fusion 4MP/6MP  
On-screen display (OSD)  
Luminance and color menu  
Measured luminance  
Indicates the actual luminance measured  
by the internal sensor. This is a read-only  
value. It is expressed in cd/m2.  
Luminance target  
Color target  
Allows to manually adjust the luminance  
target. See note below.  
Allows to select from a list of factory-  
defined and user-defined color targets.  
Viewing mode  
Allows to select the viewing mode:  
Diagnostic or Text. In text mode the  
luminance is approximately half of the  
luminance in diagnostic mode. This is  
intended for use with office applications  
such as word processing. You can also  
switch between both modes by touching  
the up and down keys at the same  
time while the OSD is not on the screen.  
Color definition  
Jumps to the color definition submenu,  
which allows to change the definition of  
the user-defined color targets.  
When you change the luminance target, the display will adjust  
its backlight to reach the target. This can be seen in the  
Measured luminance line.  
When the luminance target cannot be reached, e.g., due to  
aging of the backlight, the Measured luminance line changes  
to Minimum value reached or Maximum value reached.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Coronis Fusion 4MP/6MP 29  
   
COLOR DEFINITION SUBMENU  
A color target is defined by its luminance target and color temperature.  
The color definition submenu allows to edit these values.  
Color name  
Indicates the color target you are editing.  
You can select another color target to edit  
by using the Up and Down touch keys  
or by rotating the control wheel.  
Luminance target  
Defined in  
Allows to select the luminance target that  
corresponds to the color target.  
Indicates how the color temperature is  
defined: in (x,y) co-ordinates or Kelvin (K).  
You can switch between both by using the  
Up and Down touch keys or by  
rotating the control wheel.  
This function is available only when the  
User color target is selected.  
x
Allows to define the x co-ordinate of the  
color temperature that corresponds to the  
color target.  
This function is available only when the  
color target is defined in (x,y) co-ordinates.  
This function is not available when the  
Native White color target is selected.  
y
Allows to define the y co-ordinate of the  
color temperature that corresponds to the  
color target.  
This function is available only when the  
color target is defined in (x,y) co-ordinates.  
This function is not available when the  
Native White color target is selected.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
30 Coronis Fusion 4MP/6MP  
Color temperature  
Allows to define the color temperature that  
corresponds to the color target in Kelvin  
units.  
This function is available only when the  
color target is defined in Kelvin units.  
This function is available only when the  
User color target is selected.  
Restore factory x, y  
values  
Allows to reset the (x,y) co-ordinates of  
the color temperature to the factory values.  
This function is not available when the  
Native White color target is selected.  
Display Function menu  
Display function  
Allows to select from a list of pre-defined  
display functions.  
If the DICOM DF is selected, a number of  
additional settings is available.  
ALC & DICOM options  
Jumps to the ALC & DICOM Options  
submenu, which allows to edit the settings  
for the DICOM display function.  
This function is available only when the  
DICOM DF is selected.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Coronis Fusion 4MP/6MP 31  
 
ALC & DICOM OPTIONS SUBMENU  
Measured ambient  
light  
Indicates the ambient light actually  
measured by the ambient light sensor on  
the front of the display. This is a read-only  
value.  
Averaged ambient  
light  
Shows the average of the measured  
ambient light since the display was  
switched on.  
When you execute the function “Measure  
ALC Value”, this value will be stored as the  
Measured ALC correction value for the  
selected reading room, unless this value is  
higher than the maximum ambient light  
determined for the selected reading room.  
Correction value  
Cont. DICOM ALC  
Shows the ambient light correction value  
that is taken into account in the calculation  
of the display function.  
Allows to switch continuous DICOM ALC  
on/off.  
When switched on, the DICOM DF is  
recalculated continuously taking the  
averaged ambient light into account.  
When switched off, the DICOM DF is  
recalculated only at the moment the  
DICOM Offset is changed (see below).  
Download from Www.Somanuals.com. All Manuals Search And Download.  
32 Coronis Fusion 4MP/6MP  
DICOM offset  
This function is available only when the  
DICOM DF is selected.When Continuous  
DICOM ALC is switched off, the DICOM DF is  
recalculated at the moment the DICOM  
Offset is changed, taking the new DICOM  
Offset into account.  
When set to “Dark Room”, the  
ambient light is not taken into  
account.  
When set to “Preset”, a preset  
ambient light value determined by  
the selected reading room is taken  
into account.  
When set to “Measured ALC”, the  
averaged ambient light, determined  
by the selected reading room, is  
taken into account.  
This function is available only when  
Continuous DICOM ALC is off.  
Reading room  
Allows to select from a pre-defined list of  
reading room types.  
You must select a reading room that  
corresponds to the type of room the  
display is installed in.  
This function is available only when  
Continuous DICOM ALC is off.  
Reading room def.  
Jumps to the reading room definition  
submenu, which allows to edit the reading  
room condition settings.  
This function is available only when  
Continuous DICOM ALC is off.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Coronis Fusion 4MP/6MP 33  
Calibration info  
Jumps to the calculation information  
submenu, which displays information  
about the values taken into account to  
recalculate the DICOM DF.  
READING ROOM DEFINITION SUBMENU  
Measured ambient  
light  
Indicates the ambient light actually  
measured by the ambient light sensor on  
the front of the display. This is a read-only  
value.  
Averaged ambient  
light  
Shows the average of the measured  
ambient light since the display was  
switched on. This is a read-only value.  
Reading room  
Indicates the reading room type you are  
editing. You can select another room to  
edit by using the Up or Down touch  
keys or rotating the control wheel.  
Max. ambient light  
This indicates the maximum ambient light  
that corresponds to the selected reading  
room.  
If the measured ambient light is higher  
than the value entered here, you should  
take measures to darken the room or  
select another reading room type.  
Preset corr. value  
This value is taken into account in the  
calculation of the DICOM DF if Continuous  
DICOM ALC is switched off and DICOM  
Offset is set to “Preset”.  
To each reading room type corresponds  
another preset value.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
34 Coronis Fusion 4MP/6MP  
Measured ALC value  
The value shown here is taken into account  
in the calculation of the DICOM DF if  
Continuous DICOM ALC is switched off and  
DICOM Offset is set to “Measured ALC”.  
When you select this line and confirm the  
selection, the actual averaged ambient  
light value is stored here, overwriting the  
former value.  
However, if the averaged ambient light  
value is higher than the maximum ambient  
light value that corresponds to the selected  
reading room, a warning is displayed  
prompting you to dim the light in the  
room, and the value will be limited to the  
maximum ambient light value.  
If you dim the light in the room you can  
execute the Measure ALC Value function  
again to enter the new, lower value.  
CALIBRATION INFORMATION SUBMENU  
Preset ambient value Shows the ambient light correction value  
/ Measured ALC  
ambient value /  
Average ambient  
value  
taken into account to calculate the DICOM  
DF, expressed in lux.  
Bright luminance  
Shows the bright luminance value taken  
into account to calculate the DICOM DF.  
Dark luminance  
Shows the dark luminance value taken into  
account to calculate the DICOM DF.  
Ambient correction  
Shows the ambient light correction value  
taken into account to calculate the DICOM  
DF, expressed in cd/m².  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Coronis Fusion 4MP/6MP 35  
Settings menu  
DPMS  
Allows to switch the display power  
management system on/off. See note  
below.  
Power LED  
Allows to switch the power LED’s on state  
on/off. The LED's orange DPMS state is not  
influenced by this setting. So, when the  
display goes into power-saving mode, the  
LED will turn orange, even if it was  
switched off by this setting  
User controls  
Allows to disable the user controls on the  
front.  
When switched off, the user cannot display  
the OSD until the user controls keycode is  
entered.  
To enable the user controls again, To  
Sound  
Allows to switch the sound on/off. When  
switched on, a short beep sounds each  
time you touch a soft touch key.  
Auto menu exit  
Allows to switch the automatic menu exit  
on/off. When switched on, the OSD is  
closed automatically when left idle for a  
certain time.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
36 Coronis Fusion 4MP/6MP  
 
OSD Orientation  
Select the desired orientation of the on-  
screen display. The default position is  
Landscape, but if you put the display in  
portrait orientation (possible but not  
supported), you should select Portrait as  
OSD orientation as well. If not, the OSD will  
not be easily readable.  
Input Switch Mode  
Select “Auto” if you wish the display to  
select the video inputs automatically. If  
only one video input is connected, the  
image will be shown in the center of the  
screen. If both video inputs are connected,  
the video signal connected to the “Left”  
input is shown on the left side of the  
screen and the signal connected to the  
“Right” connector is shown on the right  
side. If one of the video signals exceeds  
the resolution of 1640x2048, the signal is  
suppressed.  
Select “Left” to display only the signal  
connected to the “Left” video input.  
Select “Right” to display only the signal  
connected to the “Right” video input.  
Preferred Input  
Select the video resolution you wish the  
connected graphic board(s) to reproduce.  
Before changing this setting you must  
switch off the PC. After changing the  
setting, switch on the PC so that the  
graphic board changes its resolution.  
This function takes effect with plug-and-  
play graphic boards only.  
More information can also be found in the  
Getting Started manual.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Coronis Fusion 4MP/6MP 37  
Exclude I Guard zone  
Scaling  
When switched on, the video resolution of  
the preferred input is set to a value with a  
reduced number of lines. The image will be  
displayed below the I-Guard patch.  
When switched on, the display will enlarge  
(scale up) low resolution video signals if  
possible.  
The DPMS system will power down the display when the  
connected computer is left idle for a certain time.  
Barco recommends to switch DPMS on to prevent image burn-  
in (image retention) on the LCD panel..  
Information menu  
Product  
The display type  
Serial No  
SW Number  
SW Version  
Indicates the display serial number  
The order no. of the internal software  
Displays the current internal software  
version  
Display Runtime  
Indicates the total time the display has  
been operating, including the time in  
stand-by  
Backlight Runtime  
Indicates the total time the display has  
been operating, excluding the time in  
stand-by  
Download from Www.Somanuals.com. All Manuals Search And Download.  
38 Coronis Fusion 4MP/6MP  
 
Changing the viewing mode  
ABOUT VIEWING MODES  
The display contains two viewing modes. Diagnostic viewing mode  
provides the full calibrated luminance. This mode is intended for using  
the display in diagnostic applications.  
In Text mode, the luminance is approximately half of the luminance in  
Diagnostic mode. This is intended for using the display with office  
applications such as word processing.  
SWITCHING VIEWING MODES BY MEANS OF THE KEYS  
While the OSD is not on the screen, touch the Up and Down touch  
keys at the same time. As a result the display switches to another  
viewing mode. The viewing mode is indicated by a message that  
appears for a few seconds after switching the viewing mode.  
SWITCHING VIEWING MODES BY MEANS OF THE OSD MENUS  
You can also switch viewing modes in the Luminance and color menu.  
Enabling/disabling User Controls  
TO DISABLE THE USER CONTROLS:  
1 Open the OSD main menu.  
2 Open the Settings menu.  
3 Select User Controls.  
4 Select Off.  
5 Return to the Settings menu.  
6 Exit the OSD  
Now the user cannot display the OSD again until the user controls  
keycode is entered (see below).  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Coronis Fusion 4MP/6MP 39  
   
TO ENABLE THE USER CONTROLS AGAIN USING THE TOUCH KEYS:  
1 To enter the user controls keycode, the OSD must not be visible.  
2 Touch any of the soft touch keys to switch on the front  
illumination.  
3 While the illumination is on, touch the following keys in successive  
order:  
Enter , Down , Up , Down , Enter  
.
As a result, the User Controls setting in the Settings menu is switched  
on again.  
TO ENABLE THE USER CONTROLS AGAIN USING THE CONTROL WHEEL:  
1 The OSD must not be visible.  
2 Press and hold the control wheel for at least 5 seconds.  
3 Rotate the control wheel 1 step clockwise.  
4 Press the control wheel 2 times.  
5 Rotate the control wheel 1 step counterclockwise. The on-screen  
display appears.  
Steps 3 to 5 must be performed in maximum 3 seconds.  
As a result, the User Controls setting in the Settings menu is switched  
on again.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
40 Coronis Fusion 4MP/6MP  
   
Switching Power LED on/off  
TO SWITCH THE POWER LED ON/OFF:  
1 Open the OSD main menu.  
2 Open the Settings menu.  
3 Select Power LED.  
4 Select On or Off.  
5 Return to the Settings menu.  
When the display is switched in stand-by, the power LED will  
turn orange, even if the power LED setting is switched off.  
When the power LED setting is switched off, the LED will still  
turn on (green) when the display is on but receives no video  
signal.  
Switching DPMS on/off  
TO SWITCH DPMS ON/OFF:  
1 Open the main menu.  
2 Open the Settings menu.  
3 Select DPMS.  
4 Select On or Off.  
5 Return to the Settings menu.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Coronis Fusion 4MP/6MP 41  
   
Changing Display Functions  
Concepts  
The Display Function menu lets you select a display function (DF).  
When you select the DICOM DF you can define the DICOM DF to be  
recalculated taking the ambient light into account. This can be a  
continuous recalculation (“Continuous DICOM ALC” on) or single  
recalculation (“Continuous DICOM ALC” off).  
When you choose for a single recalculation, the recalculation can  
be based on an illuminance of 0 lux (“Dark Room”) a preset value  
(“Preset”) or a single, real-time measurement (“Measured ALC”).  
Both Preset and Measured ALC correction values are determined by  
the selected reading room.  
In the case of “Preset” or “Measured ALC” you must select a  
reading room that corresponds to the type of room the display is  
installed in. E.g. if the display is installed in a room where CT scans  
are observed, the “CT/MR/NM” reading room is the preferred  
selection.  
The reading room is defined by 3 parameters:  
Maximum Ambient Light defines the maximum light  
allowed in this type of room. This value can be adjusted  
within certain limits determined by the selected reading  
room.  
Preset Correction Value is the predefined correction value  
for this reading room. This value can be adjusted within  
certain limits determined by the selected reading room.  
Measured ALC Value is the averaged measured ambient  
light for this reading room. It can be changed by executing  
the “Measure ALC Value” function.  
The reading room parameters are pre-defined in the display  
according to the AAPM (American Association of Physicists in  
Medicine). However they can be changed within certain limits.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
42 Coronis Fusion 4MP/6MP  
   
Display function selection  
TO SELECT A DISPLAY FUNCTION (DF):  
1 Open the main menu.  
2 Open the Display Function menu.  
3 Select Display Function.  
4 Select the desired DF.  
5 Return to the Display Function menu.  
You can select the following DFs:  
DICOM  
Select a DICOM display function for most  
medical viewing applications.  
The DICOM function results in more visible  
grayscales in the images.  
Dynamic Gamma 2.2  
Dynamic Gamma 1.8  
These are gamma functions that are shifted  
to take into account the non-zero  
luminance of an LCD panel when driven  
with a “black” signal.  
They are especially useful in CT applications  
to improve the perception of low  
Hounsfield values.  
Native  
Test  
If you select Native, the native panel  
behavior will not be corrected.  
This is identical to the Native DF.  
Gamma 2.2  
Gamma 1.8  
Select one of these display functions in  
case the display is to replace a CRT display  
with a gamma of 1.8 or 2.2 respectively.  
User  
This is a user-definable DF.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Coronis Fusion 4MP/6MP 43  
 
ALC & DICOM Options  
Continuous DICOM ALC  
INTRODUCTION  
When you select the DICOM DF, the display offers the possibility to take  
the ambient light into account for calculating the DICOM display function.  
This function is called Continuous DICOM ALC (ALC is short for Ambient  
Light Compensation).  
When Continuous DICOM ALC is switched on, the display will  
continuously adapt the DICOM DF based on the average measured  
ambient light.  
When it is switched off, the DICOM DF will be recalculated only when  
another DICOM Offset is selected.  
TO SWITCH CONTINUOUS DICOM ALC ON/OFF:  
1 Select the DICOM display function. See To select a display  
function (DF):, on page 43  
2 In the Display Function menu, select ALC & DICOM Options.  
3 In the ALC & DICOM Options submenu, select Cont. DICOM ALC.  
4 Select On or Off.  
5 Return to the ALC & DICOM Options menu.  
The ambient light is measured by the ALC sensor at the front.  
DICOM offset  
INTRODUCTION  
When Continuous DICOM ALC is switched off, the DICOM DF will be  
recalculated only when another DICOM Offset is selected. The DICOM DF  
will be recalculated based on the new offset setting.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
44 Coronis Fusion 4MP/6MP  
   
TO SELECT A DICOM OFFSET:  
1 Select the DICOM display function. See To select a display  
function (DF):, on page 43.  
2 Switch Continuous DICOM ALC off. See To switch continuous  
3 In the ALC & DICOM Options menu, select DICOM Offset.  
4 Select an offset.  
5 Return to the ALC & DICOM Options menu.  
You can select the following offsets:  
Dark room  
The DICOM DF will be recalculated taking  
an ambient light value of 0 lux into  
account.  
This means that the measured ambient  
light does not influence the DICOM DF.  
Preset  
The DICOM DF will be recalculated taking a  
preset ambient light value into account.  
This value is determined by the selected  
Reading Room and can be adjusted  
manually within certain limits.  
Measured ALC  
The DICOM DF will be recalculated taking  
the averaged measured ambient light  
value into account that is determined by  
the selected Reading Room. This value can  
be changed by executing the “Measure ALC  
Value” function in the Reading Room  
Definition submenu.  
Reading room selection  
This function is available only when the DICOM display function is  
selected and Continuous DICOM ALC is off.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Coronis Fusion 4MP/6MP 45  
TO SELECT A READING ROOM SETTING:  
1 Select the DICOM display function. See To select a display  
function (DF):, on page 43.  
2 Switch Continuous DICOM ALC off. See To switch continuous  
3 In the ALC & DICOM Options menu, select Reading Room.  
4 Select the reading room that corresponds to the room type the  
display is installed in.  
5 Return to the menu.  
You can select the following reading room types:  
DRR CR/DR/MAMMO  
DRR CT/MR/NM  
Corresponds to light conditions in  
diagnostic reading rooms for computed  
radiology, digital radiology or  
mammography. This setting has the lowest  
maximum ambient light.  
Corresponds to light conditions in  
diagnostic reading rooms for computed  
tomography, magnetic resonance or  
nuclear medicine scans.  
Staff Office  
Corresponds to light conditions in office  
rooms.  
Clinical Viewing  
Corresponds to light conditions in  
diagnostic reading rooms for clinical  
viewing.  
Emergency Room  
Operating Room  
Corresponds to light conditions in  
emergency rooms.  
Corresponds to light conditions in operating  
rooms. This setting has the highest  
maximum ambient light.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
46 Coronis Fusion 4MP/6MP  
Reading room definition  
This function is available only when the DICOM display function is  
selected and Continuous DICOM ALC is off.  
TO MODIFY A READING ROOM DEFINITION:  
1 Select the DICOM display function. See To select a display  
function (DF):, on page 43.  
2 Switch Continuous DICOM ALC off. See To switch continuous  
3 In the ALC & DICOM Options menu, select Reading Room Def.  
4 Select the reading room definition you wish to modify. Thereto,  
first select Reading Room. Select the appropriate reading room.  
5 To modify the maximum ambient light for this reading room,  
select Max. Ambient Light and change the value.  
You can change this value within certain limits determined by the  
selected reading room.  
6 To modify the Preset correction value for this reading room, select  
Preset Corr. Value and change the value.  
You can change this value within certain limits determined by the  
selected reading room.  
7 To modify the Measured ALC Value for this reading room, select  
Measure ALC Value and execute the function.  
If the “Averaged Ambient Light” at that moment is lower than the  
defined “Maximum Ambient Light”, the averaged ambient light  
value is copied as Measured ALC value.  
If the “Averaged Ambient Light” at that moment is higher than the  
defined “Maximum Ambient Light”, the display will prompt a  
warning for the user to take measures to adapt the ambient light  
in the room to the selected reading room. Proceed as follows:  
Use Enter  
or click the control wheel. As a result the  
Measured ALC value will be limited to the maximum ambient  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Coronis Fusion 4MP/6MP 47  
light value. Use Enter  
store the value.  
or click the control wheel again to  
However, if you wish to adapt the ambient light conditions to  
the definition of the selected reading room, dim the light in  
the room so that the measured ambient light becomes lower  
that the maximum ambient light.  
Then use Enter  
or click the control wheel again to execute  
the Measure ALC Value function once more. As a result the  
actually measured ambient light value will be copied as  
Measured ALC Value. Finally use Enter  
wheel again to store the value.  
or click the control  
Download from Www.Somanuals.com. All Manuals Search And Download.  
48 Coronis Fusion 4MP/6MP  
Cleaning your display  
Front glass  
Clean the front glass using a sponge, cleaning cloth or soft tissue, lightly  
Before cleaning the front glass, unplug the power cable from  
the mains to prevent the keys from being activated  
inadvertently by sweeping over them.  
moistened with one of the following tested products:  
Flux  
Pril  
Windex Glass Plus  
Bohle glass cleaner  
Mr. Proper  
Ajax glass cleaner  
Sidolin glass cleaner  
25% ethanol solution (EtOH)  
In case none of the above cleaning products is available, use plain water.  
Take care not to damage or scratch the front glass. Be careful  
with rings or other jewelry and do not apply excessive pressure  
on the front glass.  
Do not apply or spray liquid directly to the front glass as excess  
liquid may cause damage to internal electronics. Instead, apply  
the liquid to the cleaning cloth.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Coronis Fusion 4MP/6MP 49  
   
Do not use following products:  
Alcohol/solvents at higher concentration > 5%  
Strong alkalis lye, strong solvents  
Acid  
Detergents with fluoride  
Detergents with ammonia  
Detergents with abrasives  
Steel wool  
Sponge with abrasives  
Steel blades  
Cloth with steel thread  
Cabinet  
1 Clean the cabinet using a soft cotton cloth, lightly moistened with  
a recognized cleaning product for medical equipment.  
The cabinet has been tested for resistance to the following  
products:  
Cidex, Betadine  
Alcohol (Isopropyl and Ethyl)  
Ammonia-based cleaners (Windex)  
Aquasonic Gel  
In case none of the above cleaning products is available, use plain  
water.  
Do not apply or spray liquid directly to the cabinet as  
excess liquid may cause damage to internal electronics.  
Instead, apply the liquid to the cleaning cloth.  
2 Repeat with water only.  
3 Wipe dry with a dry cloth.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
50 Coronis Fusion 4MP/6MP  
 
Important information  
Safety information  
GENERAL RECOMMENDATIONS  
Read the safety and operating instructions before operating the display.  
Retain safety and operating instructions for future reference.  
Adhere to all warnings on the display and in the operating instructions  
manual.  
Follow all instructions for operation and use.  
ELECTRICAL SHOCK  
TYPE OF PROTECTION (ELECTRICAL):  
Display with external power supply: Class I equipment  
DEGREE OF SAFETY (FLAMMABLE ANESTHETIC MIXTURE):  
Equipment not suitable for use in the presence of a flammable  
anesthetic mixture with air or with oxygen or nitrous oxide.  
NON-PATIENT CARE EQUIPMENT  
Equipment primarily for use in a health care facility that is intended for  
use where contact with a patient is unlikely.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Coronis Fusion 4MP/6MP 51  
   
POWER CONNECTION - DISPLAY WITH EXTERNAL 24 VDC POWER  
SUPPLY  
Power requirements: The display must be powered using the  
delivered medical approved 24 VDC SELV power supply.  
The medical approved DC power supply must be powered by the  
AC mains voltage (protective earth terminal).  
TRANSIENT OVER-VOLTAGE  
If the equipment is not used for a long time, disconnect it from the AC  
inlet to avoid damage by transient over-voltage.  
To fully disengage the power to the unit, please disconnect the power  
cord from the AC inlet.  
POWER CORDS:  
Utilize a UL-listed detachable power cord, 3-wire, type SJ or  
equivalent, 18 AWG min., rated 300 V min., provided with a  
hospital-grade type plug 5-15P configuration for 120V application,  
or 6-15P for 240V application.  
Do not overload wall outlets and extension cords as this may result  
in fire or electric shock.  
Mains lead protection (U.S.: Power cord): Power cords should be  
routed so that they are not likely to be walked upon or pinched by  
items placed upon or against them, paying particular attention to  
cords at plugs and receptacles.  
WATER AND MOISTURE  
Never expose the display to rain or moisture.  
Never use the display near water - e.g. near a bathtub, washbasin,  
swimming pool, kitchen sink, laundry tub or in a wet basement.  
VENTILATION  
Do not cover or block the ventilation openings in the cover of the set.  
When installing the display in a cupboard or another closed location,  
Download from Www.Somanuals.com. All Manuals Search And Download.  
52 Coronis Fusion 4MP/6MP  
heed the necessary space between the set and the sides of the  
cupboard.  
INSTALLATION  
Place the display on a flat, solid and stable surface that can support the  
weight of at least 3 displays. If you use an unstable cart or stand, the  
display may fall, causing serious injury to a child or adult, and serious  
damage to the equipment.  
THIS APPARATUS CONFORMS TO:  
CE0120 (MDD 93/42/EEC; A1:2007/47/EC class IIb product), CE - 2004/  
108/EC, CE - 93/42/EEC; A1:2007/47/EC class II b, IEC 60950-1:2005 +  
A1:2009 (2ND EDITION), IEC 60601-1 2ND ED:1988 + A1:1991 + A2:1995,  
UL 60601-1 1ST EDITION, CAN/CSA-C22.2 NO. 601.1-M90:2005, IEC  
60601-1:2005 + A1:2012, ANSI/AAMI ES 60601-1:2005 + C1:2009 +  
A1:2012, CAN/CSAC22.2 No. 60601-1-08:2008, DEMKO - EN 60601-  
1:2006, EN 60601-1-2:2007, CCC - GB9254-2008 + GB4943-2001 +  
GB17625.1-2003, KC, VCCI, FCC class B, ICES-001 Level B, FDA 510K,  
RoHS.  
NATIONAL SCANDINAVIAN DEVIATIONS FOR CL. 1.7.2:  
Finland: "Laite on liitettävä suojamaadoituskoskettimilla varustettuun  
pistorasiaan"  
Norway: "Apparatet må tilkoples jordet stikkontakt"  
Sweden: "Apparaten skall anslutas till jordat uttag"  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Coronis Fusion 4MP/6MP 53  
Environmental information  
DISPOSAL INFORMATION (WASTE ELECTRICAL AND ELECTRONIC  
EQUIPMENT)  
This symbol on the product indicates that, under European  
Directive 2002/96/EC governing waste from electrical and  
electronic equipment, this product must not be disposed of  
with other municipal waste. Please dispose of your waste  
equipment by handing it over to a designated collection point  
for the recycling of waste electrical and electronic equipment. To prevent  
possible harm to the environment or human health from uncontrolled  
waste disposal, please separate these items from other types of waste  
and recycle them responsibly to promote the sustainable reuse of  
material resources.  
For more information about recycling of this product, please contact your  
local city office, your municipal waste disposal service or the shop where  
you purchased the product.  
MERCURY NOTICE  
This Barco product consists of materials that may contain mercury, which  
must be recycled or disposed of in accordance with local, state, or  
country laws:  
Within this system, the backlight lamps in the monitor display  
contain mercury  
TURKEY ROHS COMPLIANCE  
Republic of Turkey: In conformity with the EEE Regulation  
Türkiye Cumhuriyeti: EEE Yönetmeliğine Uygundur  
Download from Www.Somanuals.com. All Manuals Search And Download.  
54 Coronis Fusion 4MP/6MP  
 
中国大陆 ROHS (CHINESE MAINLAND ROHS)  
根据中国大陆 《电子信息产品污染控制管理办法也称为中国大  
陆 RoHS以下部分列出了 Barco 产品中可能包含的有毒和 / 或有  
害物质的名称和含量。中国大陆 RoHS 指令包含在中国信息产业部  
MCV 标准:“ 电子信息产品中有毒物质的限量要求 ” 中。  
According to the "China Administration on Control of Pollution Caused by  
Electronic Information Products" (Also called RoHS of Chinese Mainland),  
the table below lists the names and contents of toxic and/or hazardous  
substances that Barco's product may contain. The RoHS of Chinese  
Mainland is included in the MCV standard of the Ministry of Information  
Industry of China, in the section "Limit Requirements of toxic substances  
in Electronic Information Products".  
零件项目 ( 名称 )  
有毒有害物质或元素  
Component Name  
Hazardous Substances or Elements  
六价铬  
多溴联苯  
多溴二苯醚  
(Pb)  
(Hg)  
(Cd)  
(Cr6+)  
(PBB)  
(PBDE)  
印制电路配件  
Printed Circuit  
X
O
O
O
O
O
Assemblies  
液晶面板  
LCD panel  
X
X
X
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
外接电 ( 线 ) 缆  
External cables  
內部线路  
Internal wiring  
O
O
O
O
O
金属外壳  
Metal enclosure  
塑胶外壳  
Plastic enclosure  
散热片 ( 器 )  
Heatsinks  
风扇  
Fan  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Coronis Fusion 4MP/6MP 55  
电源供应器  
Power supply unit  
X
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
O
文件说明书  
Paper manuals  
光盘说明书  
CD manual  
O: 表示该有毒有害物质在该部件所有均质材料中的含量均在 SJ/T 11363-2006 标准规定的限  
量要求以下 .  
O: Indicates that this toxic or hazardous substance contained in all of the homogeneous  
materials for this part is below the limit requirement in SJ/T11363-2006.  
X: 表示该有毒有害物质至少在该部件的某一均质材料中的含量超出 SJ/T 11363-2006 标准规  
定的限量要求 .  
X: Indicates that this toxic or hazardous substance contained in at least one of the  
homogeneous materials used for this part is above the limit requirement in SJ/T11363  
2006.  
在中国大陆销售的相应电子信息产品 (EIP)都必须遵照中国大陆  
《电子信息产品污染控制标识要求》标准贴上环保使用期限  
(EFUP)标签。Barco 产品所采用的 EFUP 标签 (请参阅实例,徽  
标内部的编号使用于制定产品)基于中国大陆的 《电子信息产品  
环保使用期限通则》标准。  
All Electronic Information Products (EIP) that are sold within Chinese  
Mainland must comply with the "Electronic Information Products  
Pollution Control Labeling Standard" of Chinese Mainland, marked with  
the Environmental Friendly Use Period (EFUP) logo. The number inside  
the EFUP logo that Barco uses (please refer to the photo) is based on the  
"Standard of Electronic Information Products Environmental Friendly Use  
Period" of Chinese Mainland.  
10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
56 Coronis Fusion 4MP/6MP  
Regulatory compliance information  
INDICATIONS FOR USE: CORONIS FUSION 4MP  
The Coronis Fusion 4MP (MDCC-4130) display is a display device capable  
of displaying two 2MP display images electronically fused together on  
the one panel. It is intended to be used in displaying and viewing digital  
images for review and analysis by trained medical practitioners. It  
provides a screen resolution of 2560x 1600 pixels at a pixel pitch of  
0.2505mm. This provides the equivalent resolution of 2 x 2MP displays.  
The Coronis Fusion 4MP (MDCC-4130) display is not intended for primary  
diagnostic review of Mammography images.  
Caution (USA): Federal law restricts this device to sale by or on the order  
of a physician or a practitioner trained on its use.  
INDICATIONS FOR USE: CORONIS FUSION 6MP  
The Coronis Fusion 6MP DL (MDCC-6130) display is a display device  
capable of displaying two 3MP display images electronically fused  
together on the one panel. It is intended to be used in displaying and  
viewing digital images for review and analysis by trained medical  
practitioners. It provides a screen resolution of 3280 x 2048 pixels at a  
pixel pitch of 0.1995mm. This provides the equivalent resolution of 2 x  
3MP displays.  
The Coronis Fusion 6MP DL (MDCC-6130) display is not intended for  
primary diagnostic review of Mammography images.  
Caution (USA): Federal law restricts this device to sale by or on the order  
of a physician or a practitioner trained on its use.  
FCC CLASS B  
This device complies with Part 15 of the FCC Rules. Operation is subject  
to the following two conditions: (1) this device may not cause harmful  
interference, and (2) this device must accept any interference received,  
including interference that may cause undesired operation.  
This equipment has been tested and found to comply with the limits for  
a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Coronis Fusion 4MP/6MP 57  
 
are designed to provide reasonable protection against harmful  
interference in a residential installation. This equipment generates, uses  
and can radiate radio frequency energy and, if not installed and used in  
accordance with the instructions, may cause harmful interference to  
radio communications. However, there is no guarantee that interference  
will not occur in a particular installation. If this equipment does cause  
harmful interference to radio or television reception, which can be  
determined by turning the equipment off and on, the user is encouraged  
to try to correct the interference by one or more of the following  
measures:  
Reorient or relocate the receiving antenna.  
Increase the separation between the equipment and receiver.  
Connect the equipment into an outlet on a circuit different from  
that to which the receiver is connected.  
Consult the dealer or an experienced radio/TV technician for help.  
CANADIAN NOTICE  
This ISM device complies with Canadian ICES-001.  
Cet appareil ISM est conforme à la norme NMB-001 du Canada.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
58 Coronis Fusion 4MP/6MP  
EMC notice  
No specific requirement on the use of external cables or other  
accessories except power supply.  
With the installation of the device, use only the delivered power supply  
or a spare part provided by the legal manufacturer. Using another can  
result in a decrease of the immunity level of the device.  
ELECTROMAGNETIC EMISSIONS  
The CORONIS FUSION 4MP/6MP display is intended for use in the  
electromagnetic environment specified below.  
The customer or the user of the CORONIS FUSION 4MP/6MP display should  
assure that it is used in such an environment.  
Emissions test  
RF emissions  
CISPR 11  
Compliance  
Electromagnetic environment – guidance  
Group 1  
The CORONIS FUSION 4MP/6MP display uses  
RF energy only for its internal function.  
Therefore, its RF emissions are very low and  
are not likely to cause any interference in  
nearby electronic equipment  
RF emissions  
CISPR 11  
Class B  
Class D  
The CORONIS FUSION 4MP/6MP display is  
suitable for use in all establishments,  
including domestic establishments and  
those directly connected to the public low-  
voltage power supply network that supplies  
buildings used for domestic purposes.  
Harmonic emissions  
IEC 61000-3-2  
Voltage fluctuations/  
flicker emissions  
Complies  
IEC 61000-3-3  
This CORONIS FUSION 4MP/6MP display complies with appropriate medical  
EMC standards on emissions to, and interference from surrounding  
equipment. Operation is subject to the following two conditions: (1) this  
device may not cause harmful interference, and (2) this device must  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Coronis Fusion 4MP/6MP 59  
 
accept any interference received, including interference that may cause  
undesired operation.  
Interference can be determined by turning the equipment off and on.  
If this equipment does cause harmful interference to, or suffer from  
harmful interference of, surrounding equipment, the user is encouraged  
to try to correct the interference by one or more of the following  
measures:  
Reorient or relocate the receiving antenna or equipment.  
Increase the separation between the equipment and receiver.  
Connect the equipment into an outlet on a circuit different from  
that to which the receiver is connected.  
Consult the dealer or an experienced technician for help  
Download from Www.Somanuals.com. All Manuals Search And Download.  
60 Coronis Fusion 4MP/6MP  
ELECTROMAGNETIC IMMUNITY  
The CORONIS FUSION 4MP/6MP display is intended for use in the  
electromagnetic environment specified below. The customer or the user  
of the CORONIS FUSION 4MP/6MP display should assure that it is used in  
such an environment.  
Immunity test  
IEC 60601  
Test levels  
Compliance  
level  
Electromagnetic  
environment –  
guidance  
Electrostatic  
discharge (ESD)  
± 6kV contact  
± 8kV air  
± 6kV contact  
± 8kV air  
Floors should be wood,  
concrete or ceramic tile.  
If floors are covered with  
synthetic material, the  
relative humidity should  
be at least 30%  
IEC 61000-4-2  
Electrical fast  
transient/burst  
± 2kV for power  
supply lines  
± 2kV for power  
supply lines  
Mains power quality  
should be that of a  
typical commercial or  
hospital environment  
IEC 61000-4-4  
± 1kV for input/  
output lines  
± 1kV for input/  
output lines  
Surge  
± 1 kV line(s) to  
line(s)  
± 1 kV line(s) to  
line(s)  
Mains power quality  
should be that of a  
typical commercial or  
hospital environment  
IEC61000-4-5  
± 2 kV line(s) to  
earth  
± 2 kV line(s) to  
earth  
Voltage dips,  
short  
< 5% UT (> 95%  
dip in UT) for 0.5  
cycle  
< 5% UT (> 95%  
dip in UT) for 0.5  
cycle  
Mains power quality  
should by that of a  
typical commercial or  
hospital environment. If  
interruptions  
and voltage  
variations on  
power supply  
input lines  
40% UT (60%  
dip in UT) for 5  
cycles  
40% UT (60% dip  
in UT) for 5 cycles  
the user of the CORONIS  
FUSION 4MP/6MP display  
requires continued  
operation during power  
mains interruptions, it is  
recommended that the  
70% UT (30% dip  
in UT) for 25  
cycles  
IEC 61000-4-11  
70% UT (30%  
dip in UT) for 25  
cycles  
CORONIS FUSION 4MP/6MP  
< 5% UT (>95%  
dip in UT) for 5s  
< 5% UT (>95%  
dip in UT) for 5s  
display be powered from  
an uninterruptible power  
supply or a battery.  
UT is the a.c. mains voltage prior to application of the test level  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Coronis Fusion 4MP/6MP 61  
Immunity  
test  
IEC 60601  
test level  
Level  
Electromagnetic environment –  
guidance  
Power  
3 A/m  
Not  
applic  
able  
Power frequency magnetic fields should be  
at levels characteristic of a typical location  
in a typical commercial or hospital  
environment.  
frequency  
(50/60 Hz)  
magnetic field  
IEC 61000-4-8  
Conducted RF  
IEC 61000-4-6  
3 Vrms  
3 V  
Portable and mobile RF communications  
equipment should be used no closer to any  
part of the CORONIS FUSION 4MP/6MP  
display, including cables, than the  
150 kHz to  
80 MHz  
recommended separation distance  
calculated from the equation applicable to  
the frequency of the transmitter.  
3 V/  
m
Radiated RF  
3 V/m  
IEC 61000-4-3  
80 MHz to  
2.5 GHz  
Recommended separation distance  
d 1.2  
d 1.2  
d 2.3  
P
P
P
80 MHz to 800 MHz  
800 MHz to 2.5 GHz  
Where P is the maximum output power  
rating of the transmitter in watts (W)  
according to the transmitter manufacturer  
and d is the recommended separation  
distance in meters (m).  
Field strengths from fixed RF transmitters,  
as determined by an electromagnetic site  
survey,a should be less than the  
compliance level in each frequency range.b  
Interference may occur in the vicinity of  
equipment marked with symbol:  
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)  
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV  
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic  
environment due to fixed RF transmitters, an electromagnetic site survey should be  
considered. If the measured field strength in the location in which the CORONIS FUSION 4MP/  
Download from Www.Somanuals.com. All Manuals Search And Download.  
62 Coronis Fusion 4MP/6MP  
6MP display is used exceeds the applicable RF compliance level above, the CORONIS FUSION  
4MP/6MP display should be observed to verify normal operation. If abnormal performance  
is observed, additional measures may be necessary, such as re-orienting or relocating the  
CORONIS FUSION 4MP/6MP display.  
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.  
At 80 MHz and 800 MHz, the higher frequency range applies.  
These guidelines may not apply in all situations.  
Electromagnetic propagation is affected by absorption and  
reflection from structures, objects and people.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Coronis Fusion 4MP/6MP 63  
RECOMMENDED SEPARATION DISTANCE  
The CORONIS FUSION 4MP/6MP display is intended for use in an  
electromagnetic environment in which radiated RF disturbances are  
controlled. The customer of the user of the CORONIS FUSION 4MP/6MP  
display can help prevent electromagnetic interference by maintaining a  
minimum distance between portable and mobile RF communications  
equipment (transmitters) and the CORONIS FUSION 4MP/6MP display as  
recommended below,  
according to the maximum output power of the communications  
equipment.  
Rated  
Separation distance according to frequency of transmitter  
maximum  
output power  
of transmitter  
150kHz to 80MHz  
80MHz to 800MHz  
800MHz to 2.5GHz  
W
d 1.2  
P
d 1.2  
P
d 2.3 P  
0.01  
0.1  
1
0.12  
0.38  
1.2  
0.12  
0.38  
1.2  
0.23  
0.73  
2.3  
10  
3.8  
3.8  
7.3  
100  
12  
12  
23  
For transmitters rated at a maximum output power not listed above, the recommended  
separation distance d in meters (m) can be estimated using the equation applicable to the  
frequency of the transmitter. Where P is the maximum output power rating of the  
transmitter in watts (W) according to the transmitter manufacturer.  
At 80MHz and 800MHz, the separation distance for the higher  
frequency range applies.  
These guidelines may not apply in all situations.  
Electromagnetic propagation is affected by absorption and  
reflection form structures, object and people.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
64 Coronis Fusion 4MP/6MP  
Explanation of symbols  
On the display or power supply, you may find the following symbols  
(nonrestrictive list):  
Indicates compliance to the essential  
requirements of the Directive 93/42/EEC  
0120  
Indicates compliance with Part 15 of the  
FCC rules (Class A or Class B)  
Indicates the display is approved according  
to the UL regulations  
or  
Indicates the display is approved according  
to the c-UL regulations  
or  
Indicates the display is approved according  
to the DEMKO regulations  
Indicates the display is approved according  
S&E  
to the CCC regulations  
Indicates the display is approved according  
to the VCCI regulations  
Indicates the USB connectors on the display  
Indicates the manufacturing date  
Indicates the temperature limitations for  
the display to operate within specs  
Indicates the display serial no.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Coronis Fusion 4MP/6MP 65  
 
Consult the operating instructions  
Indicates this apparatus must not be  
thrown in the trash but must be recycled,  
according to the European WEEE (Waste  
Electrical and Electronic Equipment)  
directive  
Legal disclaimer  
DISCLAIMER NOTICE  
Although every attempt has been made to achieve technical accuracy in  
this document, we assume no responsibility for errors that may be  
found. Our goal is to provide you with the most accurate and usable  
documentation possible; if you discover errors, please let us know.  
Barco software products are the property of Barco. They are distributed  
under copyright by Barco N.V. or Barco, Inc., for use only under the  
specific terms of a software license agreement between Barco N.V. or  
Barco Inc. and the licensee. No other use, duplication, or disclosure of a  
Barco software product, in any form, is authorized.  
The specifications of Barco products are subject to change without  
notice.  
TRADEMARKS  
All trademarks and registered trademarks are property of their  
respective owners.  
COPYRIGHT NOTICE  
This document is copyrighted. All rights are reserved. Neither this  
document, nor any part of it, may be reproduced or copied in any form  
or by any means - graphical, electronic, or mechanical including  
Download from Www.Somanuals.com. All Manuals Search And Download.  
66 Coronis Fusion 4MP/6MP  
 
photocopying, taping or information storage and retrieval systems -  
without written permission of Barco.  
© 2013 Barco N.V. All rights reserved.  
PATENT INFORMATION  
This product is covered under the following intellectual property rights:  
US Patent RE43,707  
US Patent 7,038,186  
US Patent 7,166,829  
US Patent 6,950,098  
European Patent 1 274 066.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Coronis Fusion 4MP/6MP 67  
Technical specifications  
CORONIS FUSION 4MP  
Product acronym  
Screen technology  
MDCC-4130  
TFT AM Color LCD Dual Domain IPS-Pro  
758.36 mm (29.9")  
Active screen size  
(diagonal)  
Active screen size (H x V)  
Aspect ratio (H:V)  
Resolution  
641.208 x 404.808 mm (25.25 x 15.94")  
16:10  
native 4MP (2560 x 1600)  
configurable to 2 x 2MP+ (1280 x 1600)  
configurable to 2 x 2MP (1200 x 1600)  
Pixel pitch  
0.2505 mm  
Yes  
Color imaging  
Gray imaging  
Color support  
Viewing angle (H, V)  
Yes  
30 bit  
170°  
Uniform Luminance  
Technology (ULT)  
Yes  
Per Pixel Uniformity (PPU)  
No  
Ambient Light  
Yes  
Compensation (ALC)  
Backlight Output  
Yes (1x)  
Stabilization (BLOS)  
I-Guard  
Yes  
Maximum luminance  
1000 cd/m²  
500 cd/m²  
DICOM calibrated  
luminance (ULT off)  
Contrast ratio (typical)*  
1100:1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
68 Coronis Fusion 4MP/6MP  
 
Response time (Tr + Tf)  
Scanning frequency (H; V)  
Housing color  
20 ms  
30-150 kHz; 15-80 Hz  
Black/Silver  
Video input signals  
Video inout terminals  
USB ports  
DVI-D Dual Link (2x)  
NA  
1 upstream (endpoint), 3 downstream  
USB standard  
2.0  
Power requirements  
(nominal)  
100-240V  
Power consumption  
(nominal)  
135W  
Power save mode  
Power management  
Dot clock  
Yes  
DVI-DMPM  
280 MHz  
OSD languages  
English, French, German, Spanish, Italian  
Dimensions with stand (W  
x H x D)  
Portrait: NA  
Landscape: 731 x 580~676 x 265 mm  
Dimensions w/o stand (W  
x H x D)  
Portrait: NA  
Landscape: 731 x 485 x 141 mm  
Dimensions packaged (W x  
H x D)  
869 x 764 x 400 mm  
Net weight with stand  
Net weight w/o stand  
28.1 kg  
21.6 kg  
37.6 kg  
Net weight packaged with  
stand  
Net weight packaged w/o  
stand  
NA  
Height adjustment range  
Tilt  
96 mm  
-5° / +25°  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Coronis Fusion 4MP/6MP 69  
Swivel  
-45° / +45°  
Pivot  
NA  
Mounting standard  
Screen protection  
Recommended modalities  
Certifications  
VESA (100 mm)  
Protective, non-reflective glass cover  
CT, MR, US, DR, CR, NM, Film  
CE0120 (MDD 93/42/EEC; A1:2007/47/EC class IIb  
product), CE - 2004/108/EC, CE - 93/42/EEC;  
A1:2007/47/EC class II b, IEC 60950-1:2005 +  
A1:2009 (2ND EDITION), IEC 60601-1 2ND ED:1988 +  
A1:1991 + A2:1995, UL 60601-1 1ST EDITION, CAN/  
CSA-C22.2 NO. 601.1-M90:2005, IEC 60601-1:2005 +  
A1:2012, ANSI/AAMI ES 60601-1:2005 + C1:2009 +  
A1:2012, CAN/CSAC22.2 No. 60601-1-08:2008,  
DEMKO - EN 60601-1:2006, EN 60601-1-2:2007, CCC -  
GB9254-2008 + GB4943-2001 + GB17625.1-2003, KC,  
VCCI, FCC class B, ICES-001 Level B, FDA 510K, RoHS  
Supplied accessories  
Getting Started Guide  
Quick-Installation Sheet  
Video cables (2 x DVI Dual Link)  
Main cables (UK, European (CEBEC/KEMA), USA (UL/  
CSA; adaptor plug NEMA 5-15P), Chinese (CCC))  
USB 2.0 cable  
External power supply  
Optional accessories  
QA software  
NA  
MediCal QAWeb  
Unit per pallet  
NA  
Pallet dimensions  
Warranty  
NA  
5 years  
Operating temperature  
Storage temperature  
Operating humidity  
Storage humidity  
Operating altitude  
0°C to 35°C (15°C to 30°C within specs)  
-20°C to 60°C  
8% - 80% (non-condensing)  
5% - 95% (non-condensing)  
3000 m  
* In dark reading room conditions (0 lux)  
Download from Www.Somanuals.com. All Manuals Search And Download.  
70 Coronis Fusion 4MP/6MP  
CORONIS FUSION 6MP  
Product acronym  
Screen technology  
MDCC-6130 DL  
TFT AM Color LCD Dual Domain IPS-Pro  
772 mm (30.4")  
Active screen size  
(diagonal)  
Active screen size (H x V)  
Aspect ratio (H:V)  
Resolution  
654 x 409 mm (25.8 x 16.1")  
16:10  
native 6MP (3280 x 2048)  
configurable to 2 x 3MP+ (1640 X 2048)  
configurable to 2 x 3MP (1536 X 2048)  
Pixel pitch  
0.1995 mm  
Yes  
Color imaging  
Gray imaging  
Color support  
Viewing angle (H, V)  
Yes  
30 bit  
178°  
Uniform Luminance  
Technology (ULT)  
Yes  
Per Pixel Uniformity (PPU)  
No  
Ambient Light  
Yes  
Compensation (ALC)  
Backlight Output  
Yes (2x)  
Stabilization (BLOS)  
I-Guard  
Yes  
Maximum luminance*  
800 cd/m²  
500 cd/m²  
DICOM calibrated  
luminance (ULT off)*  
Contrast ratio (typical)*  
Response time (Tr + Tf)  
1000:1  
18 ms  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Coronis Fusion 4MP/6MP 71  
Scanning frequency (H; V)  
Housing color  
30-150 kHz; 15-80 Hz  
Black/Silver  
Video input signals  
Video inout terminals  
USB ports  
DVI-D Dual Link (2x)  
NA  
1 upstream (endpoint), 3 downstream  
USB standard  
2.0  
Power requirements  
(nominal)  
100-240V  
Power consumption  
(nominal)  
166W  
Power save mode  
Power management  
Dot clock  
Yes  
DVI-DMPM  
280 MHz  
OSD languages  
English, French, German, Spanish, Italian  
Dimensions with stand (W  
x H x D)  
Portrait: NA  
Landscape: 731 x 580~676 x 265 mm  
Dimensions w/o stand (W  
x H x D)  
Portrait: NA  
Landscape: 731 x 485 x 141 mm  
Dimensions packaged (W x  
H x D)  
869 x 764 x 400 mm  
Net weight with stand  
Net weight w/o stand  
26.9 kg  
20.4 kg  
36.1 kg  
Net weight packaged with  
stand  
Net weight packaged w/o  
stand  
NA  
Height adjustment range  
Tilt  
96 mm  
-5° / +25°  
Download from Www.Somanuals.com. All Manuals Search And Download.  
72 Coronis Fusion 4MP/6MP  
Swivel  
-45° / +45°  
Pivot  
NA  
Mounting standard  
Screen protection  
Recommended modalities  
Certifications  
VESA (100 mm)  
Protective, non-reflective glass cover  
CT, MR, US, DR, CR, NM, Film  
CE0120 (MDD 93/42/EEC; A1:2007/47/EC class IIb  
product), CE - 2004/108/EC, CE - 93/42/EEC;  
A1:2007/47/EC class II b, IEC 60950-1:2005 +  
A1:2009 (2ND EDITION), IEC 60601-1 2ND ED:1988 +  
A1:1991 + A2:1995, UL 60601-1 1ST EDITION, CAN/  
CSA-C22.2 NO. 601.1-M90:2005, IEC 60601-1:2005 +  
A1:2012, ANSI/AAMI ES 60601-1:2005 + C1:2009 +  
A1:2012, CAN/CSAC22.2 No. 60601-1-08:2008,  
DEMKO - EN 60601-1:2006, EN 60601-1-2:2007, CCC -  
GB9254-2008 + GB4943-2001 + GB17625.1-2003, KC,  
VCCI, FCC class B, ICES-001 Level B, FDA 510K, RoHS  
Supplied accessories  
Getting Started Guide  
Quick-Installation Sheet  
Video cables (2 x DVI Dual Link)  
Main cables (UK, European (CEBEC/KEMA), USA (UL/  
CSA; adaptor plug NEMA 5-15P), Chinese (CCC))  
USB 2.0 cable  
External power supply  
Optional accessories  
QA software  
NA  
MediCal QAWeb  
Unit per pallet  
NA  
Pallet dimensions  
Warranty  
NA  
5 years  
Operating temperature  
Storage temperature  
Operating humidity  
Storage humidity  
Operating altitude  
0°C to 35°C (15°C to 30°C within specs)  
-20°C to 60°C  
8% - 70% (non-condensing)  
5% - 95% (non-condensing)  
3000 m  
* In dark reading room conditions (0 lux)  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Coronis Fusion 4MP/6MP 73  
Download from Www.Somanuals.com. All Manuals Search And Download.  
74 Coronis Fusion 4MP/6MP  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Coronis Fusion 4MP/6MP 75  
K5902015-08  
February 2013  
Barco n.v.  
President Kennedypark 35  
8500 Kortrijk  
Belgium  
0120  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Aiwa Car Stereo System CDC X145 User Manual
Aiwa DVD VCR Combo HV DH1EH User Manual
Alpine Car Stereo System DVA 9965 User Manual
Amana Microwave Oven F1331E User Manual
Asus Network Card P4S333 VM User Manual
Asus Network Card PR DLSR533 User Manual
Ativa Paper Shredder DSC200C User Manual
Atmel Stereo Amplifier ATR7040 User Manual
Audiovox Stereo System 128 4874A User Manual
Badger Basket Crib 00750 User Manual