Agilent Technologies Video Gaming Accessories Agilent E3614A Serials KR83503035 User Manual

Iꢀ  
Agilent E361xA 60W BENCH SERIES DC POWER SUPPLIES  
OPERATING AND SERVICE MANUAL FOR MODELS:  
Agilent E3614A, Serials KR83503035 and above  
Agilent E3615A, Serials KR83506197 and above  
Agilent E3616A, Serials KR83502651 and above  
Agilent E3617A, Serials KR83502522 and above  
For instruments with higher Serial Numbers than above,  
a change page may be included.  
Manual Part No. 5959-5310  
April 2000  
Edition 8  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table of Contents  
SAFETY SUMMARY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2  
GENERAL INFORMATION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4  
INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4  
SAFETY REQUIREMENTS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4  
INSTRUMENT AND MANUAL IDENTIFICATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4  
OPTIONS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4  
ACCESSORY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4  
DESCRIPTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4  
SPECIFICATIONS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5  
INSTALLATION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6  
INITIAL INSPECTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6  
Mechanical Check. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6  
Electrical Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6  
INSTALLATION DATA. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6  
Location and Cooling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6  
Outline Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6  
Rack Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6  
INPUT POWER REQUIREMENTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6  
Line Voltage Option Conversion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6  
Power Cord . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7  
OPERATING INSTRUCTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-7  
INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7  
TURN-ON CHECKOUT PROCEDURE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7  
OPERATING MODES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-8  
LOCAL OPERATING MODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8  
Constant Voltage Operaton. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8  
Constant Current Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8  
Overvoltage Protection (OVP). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8  
CONNECTING LOADS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8  
OPERATION BEYOND RATED OUTPUT. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8  
REMOTE OPERATING MODES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9  
Remote Voltage Sensing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9  
Remote Analog Voltage Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9  
MULTIPLE-SUPPLY OPERATION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-10  
NORMAL PARALLEL OPERATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-10  
AUTO-PARALLEL OPERATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-10  
NORMAL SERIES OPERATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-11  
AUTO-SERIES OPERATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-12  
AUTO-TRACKING OPERATON . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-13  
LOAD CONSIDERATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-14  
PULSE LOADING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-14  
REVERSE CURRENT LOADING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-14  
OUTPUT CAPACITANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-14  
REVERSE VOLTAGE LOADING. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-14  
BATTERY CHARGING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-14  
1-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
OPTIONS  
GENERAL INFORMATION  
Options OE3 and OE9 determine which line voltage is  
selected at the factory. The standard unit is configured for 115  
Vac 10%. For information about changing the line voltage  
setting, see paragraph "INPUT POWER REQUIREMENTS",  
page 1-6.  
INTRODUCTION  
This manual describes all models in the Agilent E361xA 60W  
Bench Power Supply family and unless stated otherwise, the  
information in this manual applies to all models.  
OE3:  
OE9:  
910:  
Input power, 230 Vac 10%, 47-63 Hz  
Input power, 100 Vac 10%, 47-63 Hz  
One additional manual  
SAFETY REQUIREMENTS  
This product is a Safety Class I instrument, which means  
that it is provided with a protective earth ground terminal.  
This terminal must be connected to an ac source that has a  
3-wire ground receptacle. Review the instrument rear panel  
and this manual for safety markings and instructions before  
operating the instrument. Refer to the Safety Summary page  
at the beginning of this manual for a summary of general  
safety information. Specific safety information is located at  
the appropriate places in this manual.  
ACCESSORY  
The accessory listed below may be ordered from your local  
Agilent Technologies Sales Office either with the power sup-  
ply or separately. (Refer to the list at the rear of the manual for  
address.)  
Agilent Part No.Description  
5063-9240  
Rack Kit for mounting one or two 3 1/2" high  
supply in a standard 19" rack  
This power supply is designed to comply with the following  
safety and EMC(Electromagnetic Compatibility) require-  
ments:  
The rack mount kit is needed for rack mounting of all models  
in the Agilent E361xA power supply because these supplies  
have molded feet.  
nIEC 348: Safety Requirements for Electronic Measuring  
Apparatus  
nIEC 1010-1/EN 61010: Safety Requirements for Electrical  
Equipment for Measurement, Control, and Laboratory Use  
nCSA C22.2 No.231: Safety Requirements for Electrical and  
Electronic Measuring and Test Equipment  
nUL 1244: Electrical and Electronic Measuring and Testing  
Equipment.  
DESCRIPTION  
This power supply is suitable for either bench or rack  
mounted operation. It is a compact, well-regulated, Constant  
Voltage/Constant Current supply that will furnish full rated  
output voltage at the maximum rated output current or can be  
continuously adjusted throughout the output range. The out-  
put can be adjusted both locally from the front panel and  
remotely by changing the settings of the rear panel switches  
(See paragraph "REMOTE OPERATING MODES", page 1-9).  
The models in this family offer up to 60 watts of output power,  
with voltage up to 60 volts and current up to 6 amps as shown  
in Table 1.  
nEMC Directive 89/336/EEC: Council Directive entitled  
Approximation of the Laws of the Member States relating to  
Electromagnetic Compatibility  
nEN 55011(1991) Group 1, Class B/CISPR 11: Limits and  
nMethods of Radio Interference Characteristics of  
nIndustrial, Scientific, and Medical(ISM) Radio-Frequency  
Equipment  
nEN 50082-1(1991) /  
The front panel VOLTAGE control can be used to establish  
the voltage limit when the supply is used as a constant cur-  
rent source and the CURRENT control can be used to estab-  
lish the output current limit when the supply is used as a  
constant voltage source. The supply will automatically cross  
over from constant voltage to constant current operation and  
vice versa if the output current or voltage exceeds these pre-  
set limits.  
IEC 801-2(1991):Electrostatic Discharge Requirements  
IEC 801-3(1984):Radiated Electromagnetic Field  
Requirements  
IEC 801-4(1988):Electrical Fast Transient/Burst  
Requirements  
INSTRUMENT AND MANUAL IDENTIFICATION  
A serial number identifies your power supply. The serial  
number encodes the country of manufacture, the date of the  
latest significant design change, and a unique sequential  
number. As an illustration, a serial number beginning with  
KR306 denotes a power supply built in 1993 (3=1 993,  
4=1994, etc), 6th week manufacture in Korea(KR). The  
remaining digits of the serial number are a unique, five-digit  
number assigned sequentially.  
The front panel includes an autoranging (E3614A single-  
range) digital voltmeter and a single-range digital ammeter.  
Two 3 1/2 digit voltage and current displays accurately show  
the output voltage and current respectively. The output rat-  
ings for each model are shown in the Specifications and  
Operating Characteristics Table.  
The OVP/CC SET switch is used to check the OVP trip volt-  
age and current control set value. When pressing this switch,  
the voltage display indicates the OVP trip voltage and the cur-  
rent display indicates the current control set value.  
If the serial number on your supply differs from that shown  
on the title page of this manual,  
a
yellow MANUAL  
CHANGES sheet is supplied with this manual to explain  
the difference between your instrument and the instrument  
described by this manual. The change sheet may also con-  
tain information for correcting errors in the manual.  
The power supply has both front and rear output terminals.  
Either the positive or negative output terminal may be  
1-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
be grounded or the power supply can be operated float-  
ing at up to a maximum of 240 Volts off ground. Total out-  
put voltage to ground must not exceed 240 Vdc.  
SPECIFICATIONS  
Detailed specifications for the power supply are given in Table  
1. All specifications are at front terminals with a resistive load,  
and local sensing unless otherwise stated. Operating charac-  
teristics provide useful, but non-warranted information in the  
form of the nominal performance.  
LINE FUSE  
Line Voltage  
100/115 Vac  
230 Vac  
Fuse  
2.0 AT  
1.0 AT  
Agilent Part No.  
2110-0702  
2110-0457  
Table 1. Specifications and Operating Characteristics  
*STABILITY (OUTPUT DRIFT)  
*AC INPUT  
An internal switch permits operation from 100, 115, or 230 Vac  
lines.  
Maximum change in output for an 8 hours following a 30 minute  
warm-up under constant line, load and ambient temperature.  
Constant Voltage: Less than 0.1% plus 5 mV  
100 Vac ± 10%, 47-63 Hz, 163 VA, 125 W  
115 Vac ± 10%, 47-63 Hz, 163 VA, 125 W  
230 Vac ± 10%, 47-63 Hz, 163 VA, 125 W  
Constant Current: Less than 0.1% plus 10 mA  
LOAD TRANSIENT RESPONSE TIME  
DC OUTPUT  
Less than 50 µsec for output recovery to within 15 mV following a  
change in output current from full load to half load, or vice versa.  
Voltage and current can be programmed via front panel control or  
remote analog control over the following ranges;  
E3614A: 0 - 8 V, 0 - 6 A  
METER ACCURACY:B±(0.5% of output + 2 counts)Bat  
o
o
E3615A: 0 - 20 V, 0 - 3 A  
E3616A: 0 - 35 V, 0 - 1.7 A  
25 C ± 5 C  
E3617A: 0 - 60 V, 0 - 1 A  
METER (PROGRAMMING) RESOLUTION  
Voltage: E3614A 10 mV  
E3615A 10 mV (0 to 20 V), 100 mV (above 20 V)  
E3616A 10 mV (0 to 20 V), 100 mV (above 20 V)  
E3617A 10 mV (0 to 20 V), 100 mV (above 20 V)  
Current: E3614A 10 mA  
*OUTPUT TERMINALS  
The output terminals are provided on the front and rear panel.  
They are isolated from the chassis and either the positive or neg-  
ative terminal may be connected to the ground terminal.  
E3615A 10 mA  
E3616A 1 mA  
E3617A 1 mA  
LOAD REGULATION  
Constant Voltage - Less than 0.01% plus 2 mV for a full load to no  
load change in output current.  
Constant Current - Less than 0.01% plus 250 µA for a zero to  
maximum change in output voltage.  
*OVERLOAD PROTECTION  
A continuously acting constant current circuit protects the power  
supply for all overloads including a direct short placed across the  
terminals in constant voltage operation. The constant voltage cir-  
cuit limits the output voltage in the constant current mode of oper-  
ation.  
LINE REGULATION  
Constant Voltage - Less than 0.01% plus 2 mV for any line volt-  
age change within the input rating.  
Constant Current - Less than 0.01% plus 250 µA for any line volt-  
age change within the input rating.  
*OVERVOLTAGE PROTECTION  
Trip voltage adjustable via front panel control.  
PARD (Ripple and Noise)  
Constant Voltage: Less than 200 µV rms and 1 mV p-p  
(20 Hz-20 MHz).  
Constant Current: E3614A: Less than 5 mA rms  
E3615A: Less than 2 mA rms  
E3614A  
Range: 2.5-10 V 2.5-23 V  
E3615A  
E3616A  
2.5-39 V  
E3617A  
5-65 V  
Margin: Minimum setting above output voltage to avoid  
false tripping: 4% of output + 2 V for all models  
E3616A: Less than 500 µA rms  
E3617A: Less than 500 µA rms  
*REMOTE ANALOG VOLTAGE PROGRAMMING (25 ± 5oC)  
Remotely varied voltage from 0 to 10 V provides zero to maxi-  
mum rated output voltage or current.  
Voltage: Linearity 0.5% Current: Linearity 0.5%  
The programming inputs are protected against input voltages up  
to ±40 V.  
OPERATING TEMPERATURE RANGE  
o
0 to 40 C for full rated output. Maximum current is derated 1%  
o
o
per degree C at 40 C-55 C.  
*TEMPERATURE COEFFICIENT  
REMOTE SENSING  
o
Maximum change in output per C after a 30-minute warm-up.  
Meets load-regulation specification when correcting for load-lead  
drops of up to 0.5 V per lead with sense wire resistance of less  
than 0.5 ohms per sense lead and lead lengths of less than 5  
meters.  
Constant Voltage: Less than 0.02% plus 500 µV.  
Constant Current: E3614A: Less than 0.02% plus 3 mA  
E3615A: Less than 0.02% plus 1.5 mA  
E3616A: Less than 0.02% plus 1 mA  
E3617A: Less than 0.02% plus 0.5 mA  
1-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table 1. Specifications and Operating Characteristics (Cont’d)  
*REMOTE PROGRAMMING SPEED  
DC ISOLATION  
Maximum time required for output voltage to change from initial  
value to within a tolerance band (0.1%) of the newly programmed  
value following the onset of a step change in the programming  
input voltage.  
± 240 Vdc maximum between either output terminal and earth  
ground including the output voltage.  
*COOLING: Convection cooling is employed.  
*WEIGHT: 12.1 lbs/5.5 Kg net, 14.9 lbs/6.75 Kg shipping.  
* Operating Characteristics  
Full load  
3 msec  
No load  
2 msec  
6 msec  
85 msec  
200 msec  
1.6 sec  
2.2 sec  
1.8 sec  
3.2 sec  
Up:  
E3614A:  
E3615A:  
E3616A:  
E3617A:  
9 msec  
85 msec  
200 msec  
7 msec  
13 msec  
65 msec  
200 msec  
Down: E3614A:  
E3615A:  
E3616A:  
E3617A:  
instructions.  
INSTALLATION  
INITIAL INSPECTION  
Before shipment, this instrument was inspected and found to be  
free of mechanical and electrical defects. As soon as the instru-  
ment is unpacked, inspect for any damage that may have  
occurred in transit. Save all packing materials until the inspection  
is completed. If damage is found, a claim should be filed with the  
carrier. The Agilent Technologies Sales and Service office should  
be notified.  
Mechanical Check  
This check should confirm that there are no broken knobs or connec-  
tors, that the cabinet and panel surfaces are free of dents and  
scratches, and that the meter is not scratched or cracked.  
Electrical Check  
The instrument should be checked against its electrical specifi-  
cations. Paragraph "TURN-ON CHECKOUT PROCEDURE" con-  
tains a brief checkout procedure and "PERFORMANCE TEST" in  
section SERVICE INFORMATION includes an instrument perfor-  
mance check to verify proper instrument operation.  
Figure 1. Outline Diagram  
INPUT POWER REQUIREMENTS  
This power supply may be operated from nominal 100, 115, or  
230 Vac 47-63 Hertz power source. A label on the rear panel  
shows the nominal input voltage set for the unit at the factory. If  
necessary, you can convert the supply to another nominal input  
voltage by following the instructions below  
INSTALLATION DATA  
The instrument is shipped ready for bench operation. It is neces-  
sary only to connect the instrument to a source of power and it is  
ready for operation.  
Line Voltage Option Conversion  
Location and Cooling  
Line voltage conversion is accomplished by adjusting two compo-  
nents: the line select switch and the rear panel fuse F1. To con-  
vert the supply from one line voltage option to another, proceed  
as follows:  
This instrument is air cooled. Sufficient space should be allowed so  
that a free flow of cooling air can reach the sides and rear of the  
instrument when it is in operation. It should be used in an area where  
o
the ambient temperature does not exceed 40 C. Maximum current is  
o
o
o
derated 1% per C at 40 C-55 C.  
a. Disconnect power cord.  
b. Turn off the supply and remove the top cover by lifting the  
cover upwards after taking it off from both sides of the chassis  
by inserting a flat-blade screwdriver into the gap on the lower  
rear portion of the cover.  
Outline Diagram  
Figure 1 is a outline diagram showing the dimensions of the  
instrument.  
Rack Mounting  
c. Set two sections of the line voltage selector switch on the PC  
board for the desired line voltage (see Figure 2).  
d. Check the rating of the fuse F1 installed in the rear panel fuse  
holder and replace with the correct fuse if necessary. For 100  
and 115 V operation, use a normal blow 2 A fuse and for 230  
V use a time delay 1 A fuse.  
This instrument may be rack mounted in a standard 19-inch rack  
panel either by itself or alongside a similar unit. Please see  
ACCESSORY, page 1-4, for available rack mounting accesso-  
ries. Each rack-mounting kit includes complete installation  
1-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
e. Replace the cover and mark the supply clearly with a tag or  
label indicating the correct line voltage and fuse that is in  
use.  
4. DISPLAY OVP/CC SET Switch: Pressing this switch causes  
the VOLTS display to show voltage setting for overvoltage  
shutdown (trip voltage) and the AMPS display to show the  
current control set value. Setting values are either front panel  
settings or remote voltage programmed settings.  
5. OVP Adjust Screwdriver Control: While pressing the DIS-  
PLAY OVP/CC SET switch, rotating the control clock-wise  
with a small, flat-blade screwdriver increases the setting for  
overvoltage shutdown.  
6. VOLTS Display: Digital display of actual output voltage, or  
OVP shutdown setting.  
7. AMPS Display: Digital display of actual output current, or  
output-current setting.  
8. CV LED Indicator: Output voltage is regulated when lighted.  
This means the power supply is operating in the constant  
voltage mode.  
Figure 2. Line Voltage Selector (set for 115 Vac)  
9. CC LED Indicator: Output current is regulated when lighted.  
This means the power supply is operating in the constant cur-  
rent mode.  
10. OVP LED Indicator: Output is shutdown by the occurrence  
of an overvoltage when lighted. Removing the cause of over-  
voltage and turning the power off, then on, resets the power  
supply.  
Power Cord  
To protect operating personnel, the instrument should be  
grounded. This instrument is equipped with a three conductor  
power cord. The third conductor is the ground conductor and  
when the power cord is plugged into an appropriate receptacle,  
the supply is grounded.  
The power supply was shipped with a power cord for the type of  
outlet used at your location. If the appropriate cord was not  
included, contact your nearest Agilent Sales Office to obtain the  
correct cord.  
TURN-ON CHECKOUT PROCEDURE  
The following checkout procedure describes the use of the front  
panel controls and indicators illustrated in Figure 3 and ensures  
that the supply is operational:  
OPERATING INSTRUCTIONS  
_
_
LOCAL  
_
MASTER  
+
+
+
INTRODUCTION  
This section explains the operating controls and indicators and  
provides information on many operating modes possible with your  
instrument. The front panel controls and indicators are illustrated  
in Figure 3.  
M/S 1  
M/S 2  
CV  
CC  
REMOTE  
SENSE  
+S  
OUT  
-S  
CV  
CC VREF A1 A2 A3 A4 A5  
SLAVE  
Figure 4. Switch Settings of Rear-Panel Control for Turn-  
On Checkout  
a. Disconnect power cord.  
b. Check that the rear-panel switch settings are as shown in Fig-  
ure 4.  
c. Check that the rear panel label indicates that the supply is set  
to match your input line voltage (If not, refer to "Line Voltage  
Option Conversion".).  
d. Check that the fuse on the rear panel is correct for your line  
voltage.  
e. Connect the power cord and push the LINE switch to ON.  
f. While pressing OVP/CC SET switch, verify that the OVP  
shutdown is set above 8.0, 20.0, 35.0, or 60.0 Vdc for  
E3614A, E3615A, E3616A, or E3617A respectively. If not,  
turn up OVP Adjust with a small flat-blade screwdriver.  
g. Turn VOLTAGE control fully counter clockwise to ensure that  
the output of VOLTS display decreases to 0 Vdc, then fully  
clockwise to ensure that output voltage increases to the maxi-  
mum output voltage.  
Figure 3. Front-Panel Controls and Indicators  
1. LINE Switch: Pressing this switch turns the supply on, or off.  
2. VOLTAGE Control: Clockwise rotation increases output volt-  
age.  
h. While pressing OVP/CC SET switch, turn the CURRENT con-  
trol fully counter clockwise and then fully clockwise to ensure  
3. CURRENT Control: Clockwise rotation increases output cur-  
rent.  
1-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
that the current limit value can be set from zero to maximum  
rated value.  
False OVP shutdowns may occur if you set the OVP shutdown  
too close to the supply's operating voltage. Set the OVP shut-  
down voltage 4% of output +2.0 V or more above the output volt-  
age to avoid false shutdowns from load-induced transients.  
OPERATING MODES  
The setting of the rear panel switch determines the operating  
modes of the power supply. The local operating mode is set so  
the power supply senses the output voltage directly at the output  
terminals (local sensing) for operation using the front panel con-  
trols (local programming). Other operating modes are: remote  
voltage sensing and remote programming of output voltage and  
current using external voltages.  
Adjusting OVP. Follow this procedure to adjust the OVP shut-  
down voltage.  
a. With the VOLTAGE control fully counter clockwise, turn on  
the power supply.  
b. While depressing DISPLAY OVP/CC SET switch, adjust  
the OVP Adjust control to the desired OVP shutdown using  
a small, flat-blade screwdriver.  
LOCAL OPERATING MODE  
c. Follow the procedure for CC or CV operaton to set the out-  
put voltage and current  
The power supply is shipped from the factory configured in the  
local operating mode. Local operating mode requires the switch  
settings of the rear panel, as shown in Figure 4. The power sup-  
ply provides constant voltage(CV) or constant current(CC) output.  
Resetting OVP. If OVP shutdown occurs, reset the supply by  
turning power off. Wait one or more seconds, and turn power on  
again. If OVP shutdown continue to occur, check the connections  
to the load and sense terminals, and check the OVP limit setting..  
Constant Voltage Operaton  
To set up a power supply for constant voltage operation, proceed  
as follows:  
a. Turn on the power supply and adjust 10-turn VOLTAGE con-  
trol for desired output voltage (output terminals open).  
b. While depressing DISPLAY OVP/CC SET switch, adjust 10-  
turn CURRENT control for the desired current limit.  
c. With power off connect the load to the output terminals.  
d. Turn on the power supply. Verify that CV LED is lighted.  
During actual operation, if a load change causes the current  
limit to be exceeded, the power supply will automatically  
cross over to constant current mode and the output voltage  
will drop proportionately.  
Strong electrostatic discharge to power supply can make  
OVP trip and eventually crowbar the output, which can  
effectively protect output loads from the hazardous ESD  
current.  
CONNECTING LOADS  
The output of the supply is isolated from earth ground. Either out-  
put terminal may be grounded or the output can be floated up to  
240 volts off ground. Total output voltage to ground must not  
exceed 240 Vdc.  
Constant Current Operation  
To set up a power supply for constant current operation, proceed  
as follows:  
Each load should be connected to the power supply output terminals  
using separate pairs of connecting wires. This will minimize mutual  
coupling effects between loads and will retain full advantage of the  
low output impedance of the power supply. Each pair of connecting  
wires should be as short as possible and twisted or shielded to  
reduce noise pick-up. (If a shield is used, connect one end to the  
power supply ground terminal and leave the other end unconnec-  
ted.)  
a. Turn on power supply.  
b. While depressing DISPLAY OVP/CC SET switch, adjust  
CURRENT control for the desired output current.  
c. Turn up the VOLTAGE control to the desired voltage limit.  
d. With power off connect the load to the output terminal.  
e. Turn on power supply and then verify that CC LED is lighted.  
(If CV LED is lighted, choose a higher voltage limit. A voltage  
setting that is greater than the current setting multiplied by the  
load resistance in ohms is required for CC operation.) During  
actual operation, if a load change causes the voltage limit to  
be exceeded, the power supply will automatically cross over  
to constant voltage operation at the preset voltage limit and  
output current will drop proportionately.  
If load considerations require that the output power distribution  
terminals be remotely located from the power supply, then the  
power supply output terminals should be connected to the remote  
distribution terminals via a pair of twisted or shielded wires and  
each load separately connected to the remote distribution termi-  
nals. For this case, remote sensing should be used (See para-  
graph "Remote Voltage Sensing").  
Overvoltage Protection (OVP)  
Adjustable overvoltage protection guards your load against over-  
voltage. When the voltage at the output terminals increases (or is  
increased by an external source) to the OVP shutdown voltage as  
set by the OVP ADJUST control, the supply's OVP circuit dis-  
ables the output causing the output voltage and current to drop to  
zero. During OVP shutdown the OVP LED lights.  
OPERATION BEYOND RATED OUTPUT  
The output controls can adjust the voltage or current to values up  
to 5% over the rated output. Although the supply can be operated  
in the 5% overrange region without being damaged, it can not be  
guaranteed to meet all of its performance specifications in this  
region.  
1-8  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Output Noise. Any noise picked up on the sense leads will  
appear at the supply's output voltage and may degrade CV load  
regulation. Twist the sense leads to minimize the pickup of exter-  
nal noise and run them parallel and close to the load leads. In  
noisy environments, it may be necessary to shield the sense  
leads. Ground the shield at the power supply end only. Do not use  
the shield as one of the sensing conductors.  
REMOTE OPERATING MODES  
Remote operating modes discussed below are remote voltage  
sensing and remote voltage programming. You can set up the unit  
for remote operating modes by changing the settings of the rear  
panel switch and connecting the leads from the rear panel termi-  
nals to the load or the external voltage. Solid conductors of 0.75  
2
to 1.5 mm can be connected to the rear panel terminals by sim-  
ply push fitting. Thinner wires or conductors are inserted into the  
connection space after depressing the orange opening lever.  
Stability. When the supply is connected for remote sensing, it is  
possible for the impedance of the load wires and the capacitance  
of the load to form a filter, which will become part of the supply's  
CV feedback loop. The extra phase shift created by this filter can  
degrade the supply's stability and can result in poor transient  
response performance or loop stability. In extreme cases, it can  
cause oscillations. Keep the leads as short as possible and twist  
the leads of the load to eliminate the load lead inductance and  
keep the load capacitance as small as possible.The load leads  
should be of the largest diameter practical, heavy enough to limit  
the voltage drop in each lead to 0.5 volts.  
Turn off the supply while making changes to rear panel  
switch settings or connections. This avoids the possibility  
of damage to the load and OVP shutdown from unin-  
tended output.  
Remote Voltage Sensing  
Remote voltage sensing is used to maintain good regulation at  
the load and reduce the degradation of regulation that would  
occur due to the voltage drop in the leads between the power  
supply and the load. By connecting the supply for remote voltage  
sensing, voltage is sensed at the load rather than at the supply's  
output terminals. This will allow the supply to automatically com-  
pensate for the voltage drop in the load leads and improve regula-  
tion.  
The sense leads are part of the supply's programming feedback  
control loop. Accidental open-connections of sense or load leads  
during remote sensing operation have various unwanted effects.  
Provide secure, permanent connections-especially for the sense  
leads.  
_
_
LOCAL  
_
MASTER  
+
+
+
When the supply is connected for remote sensing, the OVP circuit  
senses the voltage at the sense leads and not the main output  
terminals.  
M/S1  
M/S2  
CV  
CC  
REMOTE  
SENSE  
OUT  
+S  
-S  
CV  
CC VREF A1 A2 A3 A4 A5  
SLAVE  
Remote voltage sensing compensates for a voltage drop of  
up to 0.5 V in each load, and there may be up to a 0.1 V  
drop between the output terminal and the internal sensing  
resistor, at which point the OVP circuit is connected. There-  
fore, the voltage sensed by the OVP circuit could be as  
much as 1.1 V more than the voltage being regulated at the  
load. It may be necessary to re-adjust the OVP trip voltage  
when using remote sensing.  
+
nqcf  
_
pqvg\Bv™‹•–B•‡•‡BŽ‡ƒ†•Bƒ†BŽ‘ƒ†BŽ‡ƒ†•  
CV Regulation. Notice that any voltage drop in the sense leads  
adds directly to the CV load regulation. In order to maintain the  
specified performance, keep the sense lead resistance to 0.5  
ohms per lead or less.  
Figure 5. Remote Voltage Sensing  
Remote Analog Voltage Programming  
Remote analog voltage programming permits control of the regu-  
lated output voltage or current by means of a remotely varied volt-  
age. The programming (external) voltage should not exceed 10  
volts. The stability of the programming voltages directly affects  
the stability of the output. The voltage control on the front panel is  
disabled during remote analog programming.  
Remote Sensing Connections. Remote sensing requires  
changing settings of the rear panel switch and connecting the  
load leads from + and - output terminals to the load and connect-  
ing the sense leads from the +S and -S terminals to the load as  
shown in Figure 5.  
The supply includes clamp circuits to prevent it from  
supplying more than about 120% of rated output voltage  
or current when the remote programming voltage is  
greater than 10 Vdc. Do not intentionally operate the sup-  
Observe polarity when connecting the sensing leads to  
the load.  
1-9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
ply above 100% rated output. Limit your programming  
voltage to 10 Vdc.  
MULTIPLE-SUPPLY OPERATION  
Normal parallel and auto-parallel operation provides increased out-  
put current while normal series and auto-series provides increased  
output voltage. Auto-tracking provides single control of output volt-  
age of more than one supply. You can set up the unit for multiple-  
supply operation by changing the settings of the rear panel switch  
and connecting the leads from the rear panel terminals to the load.  
Remote Programming Connections. Remote programming  
requires changing settings of the switch and connecting external  
voltages to + and - terminals of "CV" or "CC" on the rear panel.  
Any noise picked up on the programming leads will appear on the  
supply's output and may degrade regulation. To reduce noise  
pick-up, use a twisted or shielded pair of wires for programming,  
with the shield grounded at one end only. Do not use the shield as  
a conductor.  
2
Solid conductors of 0.75 to 1.5 mm can be connected to the rear  
panel terminals by simply push fitting. Thinner wires or conductors  
are inserted into the connection space after depressing the orange  
opening lever.  
Notice that it is possible to operate a power supply simulta-  
neously in the remote sensing and the remote analog program-  
ming modes.  
NORMAL PARALLEL OPERATION  
Two or more power supplies being capable of CV/CC automatic  
cross over operation can be connected in parallel to obtain a total  
output current greater than that available from one power supply.  
The total output current is the sum of the output currents of the  
individual power supplies. The output of each power supply can  
be set separately. The output voltage controls of one power sup-  
ply should be set to the desired output voltage; the other power  
supply should be set for a slightly higher output voltage. The sup-  
ply with the higher output voltage setting will deliver its constant  
current output, and drop its output voltage until it equals the out-  
put of the other supply, and the other supply will remain in con-  
stant voltage operation and only deliver that fraction of its rated  
output current which is necessary to fulfill the total load demand.  
Figure 8 shows the rear panel switch settings and terminal con-  
nections for normal parallel operation of two supplies.  
Remote Programming, Constant Voltage. Figure 6 shows the  
rear panel switch settings and terminal connections for remote-  
voltage control of output voltage. A 1 Vdc change in the remote  
programming voltage produces a change in output voltage (volt-  
age gain) as follows: E3614A: 0.8 Vdc, E3615A: 2 Vdc, E3616A:  
3.5 Vdc, E3617A: 6 Vdc  
_
_
LOCAL  
_
MASTER  
+
+
+
M/S 1  
M/S 2  
CV  
CC  
REMOTE  
SENSE  
+S  
OUT  
-S  
CV  
CC  
VREF A1 A2 A3 A4 A5  
SLAVE  
NOTE:  
POWER SUPPLY  
_
_
LOCAL  
_
MASTER  
+
+
+
See the supplementary Manual, if you are not using  
isolated programming voltage source.  
Figure 6. Remote Voltage Programming, Constant  
Voltage  
M/S 1 M/S 2  
SLAVE  
CV  
CC  
SENSE  
OUT  
+S  
-S  
CV  
CC VREF A1 A2 A3 A4 A5  
REMOTE  
Remote Programming, Constant Current. Figure 7 shows the  
rear panel switch settings and terminal connections for remote-  
voltage control of output current. A 1 Vdc change in the remote  
programming voltage produces a change in output current (cur-  
rent gain) as follows: E3614A: 0.6 Adc, E3615A: 0.3 Adc,  
E3616A: 0.17 Adc, E3617A: 0.1 Adc  
LOAD  
POWER SUPPLY  
MASTER  
_
_
LOCAL  
_
+
+
+
M/S 1 M/S 2  
SLAVE  
CV  
CC  
SENSE  
+S  
OUT  
-S  
CV  
CC VREF A1 A2 A3 A4 A5  
REMOTE  
_
_
LOCAL  
_
MASTER  
+
+
+
Figure 8. Normal Parallel Operation of Two Supplies  
AUTO-PARALLEL OPERATION  
Auto-parallel operation permits equal current sharing under all load  
conditions, and allows complete control of output current from one  
master supply. The control unit is called the master; the controlled  
units are called slaves. Normally, only supplies having the same  
model number should be connected for auto-parallel operation,  
since the supplies must have the same voltage drop across the cur-  
rent monitoring resistor at full current rating. The output current of  
each slave is approximately equal to the master's. Figure 9 and Fig-  
ure 10 show the rear panel switch settings and terminal connections  
for auto-parallel operation of two supplies and three supplies.  
M/S 1  
M/S 2  
CV  
CC  
REMOTE  
SENSE  
+S  
OUT  
-S  
CV  
CC  
VREF A1 A2 A3 A4 A5  
SLAVE  
NOTE:  
See the supplementary Manual, if you are not using  
isolated programming voltage source.  
Figure 7. Remote Voltage Programming, Constant  
Current  
Remote Programming Speed. See the table of Specifications,  
page 1-5.  
1-10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Setting Voltage and Current. Turn the slave unit's CURRENT  
control fully clockwise. Adjust the master unit's controls to set the  
desired output voltage and current. The master supply operates  
in a completely normal fashion and may be set up for either con-  
stant voltage or constant current operation as required. Verify that  
the slave is in CV operation.  
gramming according to the remote-programming instructions.  
MASTER POWER SUPPLY  
_
_
_
_
_
MASTER  
LOCAL  
+
+
+
For auto-parallel operation of two supplies, the combined output  
voltage is the same as the master unit's voltage setting, and the  
combined output current is two times the master unit's current. In  
general, for two supplies, the auto-parallel output current(Io) is  
M/S 1 M/S 2  
SLAVE  
CV  
CC  
REMOTE  
SENSE  
SENSE  
SENSE  
+S  
+S  
+S  
OUT  
-S  
CV  
CC VREF A1 A2 A3 A4 A5  
LOAD  
SLAVE POWER SUPPLY  
_
_
LOCAL  
MASTER  
+
+
+
Io = Im + Is = 2Im  
where Im = master unit's output current  
Is = slave unit's output current  
M/S 1 M/S 2  
SLAVE  
CV  
CC  
OUT  
-S  
CV  
CC VREF A1 A2 A3 A4 A5  
REMOTE  
SLAVE POWER SUPPLY  
_
_
LOCAL  
MASTER  
+
+
+
Proportional currents from auto-paralleled units require  
equal load-lead voltage drops. Connect each supply to  
the load using separate pairs of wire with length chosen  
to provide equal voltage drops from pair to pair. If this is  
not feasible, connect each supply to a pair of distribution  
terminals using equal- voltage-drop wire pairs, and then  
connect the distribution terminals to the load with a single  
pair of leads.  
M/S 1 M/S 2  
SLAVE  
CV  
CC  
OUT  
-S  
CV  
CC VREF A1 A2 A3 A4 A5  
REMOTE  
Figure 10. Auto-Parallel Operation of Three Supplies  
NORMAL SERIES OPERATION  
Series operation of two or more power supplies can be accom-  
plished up to the output isolation rating of any one supply to  
obtain a higher voltage than that available from a single supply.  
Series connected supplies can be operated with one load across  
both supplies or with a separate load for each supply. These  
power supplies have a reverse polarity diode connected across  
the output terminals so that if operated in series with other sup-  
plies, damage will not occur if the load is short-circuited or if one  
supply is turned on separately from its series partners. When this  
connection is used, the output voltage is the sum of the voltages  
of the individual supplies. Each of the individual supplies must be  
adjusted in order to obtain the total output voltage. Figure 11  
shows the rear panel switch settings and terminal connections for  
normal series operation of two supplies.  
MASTER POWER SUPPLY  
_
_
LOCAL  
_
MASTER  
+
+
+
M/S 1 M/S 2  
SLAVE  
CV  
CC  
SENSE  
OUT  
VREF  
A1 A2 A3 A4 A5  
+S  
-S  
CV  
CC  
REMOTE  
LOAD  
SLAVE POWER SUPPLY  
MASTER  
_
_
LOCAL  
_
+
+
+
M/S 1 M/S 2  
SLAVE  
CV  
CC  
SENSE  
+S  
OUT  
-S  
CV  
CC VREF A1 A2 A3 A4 A5  
REMOTE  
POWER SUPPLY  
_
_
LOCAL  
_
MASTER  
Figure 9. Auto-Parallel Operation of Two Supplies  
+
+
+
Overvoltage Protection. Adjust the desired OVP shutdown limit  
using the master unit's OVP Adjust control. Set the slave units'  
OVP limits above the master's. When a master-unit shuts down,  
the master programs the slave units to zero voltage output. If a  
slave unit shuts down, it shuts only itself down. If the required cur-  
rent is great enough, the master will switch from CV to CC opera-  
tion.  
M/S 1 M/S 2  
SLAVE  
CV  
CC  
SENSE  
OUT  
+S  
-S  
CV  
CC VREF A1 A2 A3 A4 A5  
REMOTE  
LOAD  
POWER SUPPLY  
MASTER  
_
_
LOCAL  
_
+
+
+
Remote Sensing. To remote sense with auto-parallel operation,  
connect remote-sense leads only to the master unit according to  
the remote-sensing instructions.  
M/S 1 M/S 2  
SLAVE  
CV  
CC  
SENSE  
OUT  
+S  
-S  
CV  
CC VREF A1 A2 A3 A4 A5  
Remote Analog Voltage Programming. To remote program with  
REMOTE  
auto-parallel operation, set up only the master unit for remote pro-  
Figure 11. Normal Series Operation of Two Supplies  
1-11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
above the master unit's current setting to avoid having the slave  
switch to CC operation.  
AUTO-SERIES OPERATION  
Auto-series operation permits equal or proportional voltage  
sharing, and allows control of output voltage from one master  
unit. The voltage of the slaves is determined by the setting of  
the front panel VOLTAGE control on the master and voltage  
divider resistor. The master unit must be the most positive sup-  
ply of the series. The output CURRENT controls of all series  
units are operative and the current limit is equal to the lowest  
setting. If any output CURRENT controls are set too low, auto-  
matic cross over to constant current operation will occur and the  
output voltage will drop. Figure 12 and Figure 13 show the rear  
panel switch settings and terminal connections for Auto-series  
operation of two supplies and three supplies. This mode can  
also give ±voltage tracking operation of two supplies with two  
separate loads.  
When in CC operation the combined output current is the same  
as the master unit's current setting, and when in CV operation the  
combined output voltage is the sum of the master unit's and the  
slave unit's output voltages.  
Overvoltage Protection. Set the OVP shutdown voltage in each  
unit so that it shuts down at a voltage higher than its output voltage  
during auto-series operation. When a master unit shuts down, it pro-  
grams any slave units to zero output. When a slave unit shuts down,  
it shuts down only itself (and any slaves below it in the stack). The  
master (and all slaves above the shut-down slave) continues to sup-  
ply output voltage.  
Mixed model numbers may be employed in auto-series combi-  
nation without restriction, provided that each slave is specified as  
being capable of auto-series operation. If the master supply is set  
up for constant current operation, then the master-slave combina-  
tion will act as a composite constant current source.  
MASTER POWER SUPPLY  
_
_
LOCAL  
_
MASTER  
+
+
+
M/S 1 M/S 2  
SLAVE  
CV  
CC  
SENSE  
+S  
OUT  
-S  
CV  
CC VREF A1 A2 A3 A4 A5  
REMOTE  
LOAD  
R1 R2  
SLAVE POWER SUPPLY  
MASTER  
Total output voltage to ground must not exceed 240 Vdc.  
_
_
LOCAL  
_
+
+
+
Determining Resistors. External resistors control the fraction (or  
multiple) of the master unit's voltage setting that is supplied from  
the slave unit. Notice that the percentage of the total output volt-  
age contributed by each supply is independent of the magnitude  
of the total voltage. For two units in auto-series the ratio of R1 to  
R2 is  
M/S 1 M/S 2  
SLAVE  
CV  
CC  
SENSE  
+S  
OUT  
-S  
CV  
CC VREF A1 A2 A3 A4 A5  
REMOTE  
Figure 12. Auto-Series Operation of Two Supplies  
(R1+R2)/R1 = (Vo/Vm)  
R2/R1  
= (Vs/Vm)  
MASTER POWER SUPPLY  
_
_
_
LOCAL  
MASTER  
+
+
+
Where Vo = auto-series voltage = Vs + Vm  
Vm = master unit's output voltage  
Vs = slave unit's output voltage  
M/S 1 M/S 2  
SLAVE  
CV  
CC  
REMOTE  
SENSE  
+S  
OUT  
-S  
CV  
CC VREF A1 A2 A3 A4 A5  
For example, using the E3617A as a slave unit and putting R2=50  
k(1/4 watt), then from the above equations,  
R1 = R2(Vm/Vs) = 50(Vm/Vs) kΩ  
LOAD  
R1 R2  
SLAVE POWER SUPPLY(S1)  
_
_
_
LOCAL  
MASTER  
+
+
+
In order to maintain the temperature coefficient and stability perfor-  
mance of the supply, choose stable, low noise resistors.  
M/S 1 M/S 2  
SLAVE  
CV  
CC  
REMOTE  
SENSE  
+S  
OUT  
-S  
CV  
CC VREF A1 A2 A3 A4 A5  
R3 R4  
SLAVE POWER SUPPLY(S2)  
It is recommended to connect a 0.1 µF capacitor in paral-  
lel with R2 in two supplies operation or R2 and R4 in  
three supplies operation to ensure the stable operation.  
_
_
_
LOCAL  
MASTER  
+
+
+
Setting Voltage and Current. Use the master unit's controls to  
set the desired output voltage and current. The VOLTAGE control  
of the slave unit is disabled. Turning the voltage control of the  
master unit will result in a continuous variation of the output of the  
series combination, with the contribution of the master's output  
voltage to that of the slave's voltage always remaining in the ratio  
of the external resistors. Set the CURRENT control of slave unit  
M/S 1 M/S 2  
SLAVE  
CV  
CC  
SENSE  
+S  
OUT  
-S  
CV  
CC VREF A1 A2 A3 A4 A5  
REMOTE  
R2  
R1  
R2R4  
Vo=Vm(1+  
+
)
Where Vo = Auto-Series voltage = Vm + Vs1 + Vs2  
Vm = master unit's output voltage  
R1R3  
Vs1 = slave(S1) unit's output voltage  
Vs2 = slave(S2) unit's output voltage  
Figure 13. Auto-Series Operation of Three Supplies  
1-12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Remote Sensing. To remote sense with auto-series operation,  
set SENSE switch of the master unit and set SENSE switch of the  
slave unit to remote.  
Remote Analog Programming. To simultaneously remote pro-  
gram both units' output voltages, set up only the master unit for  
remote voltage programming according to the remote program-  
ming instructions. To vary the fraction of the output voltage contri-  
bution by the slave unit, connect a variable resistor in place of R2  
in two units operation. To independently remote program each  
unit's output current setting, set up each unit for remote control of  
output current according to the instructions under "Remote Pro-  
gramming, Constant Current" paragraph.  
Remote Analog Voltage Programming. To remote analog pro-  
gram with auto-series operation, connect program (external) volt-  
ages to the "CV" or "CC"" terminal of the master unit and set "CV"  
or "CC" switch of the master unit to remote.  
AUTO-TRACKING OPERATON  
Auto-tracking operation of power supplies is similar to auto-series  
operation except that the master and slave supplies have the  
same output polarity with respect to a common bus or ground.  
This operation is useful where simultaneous turn-up, turn-down or  
proportional control of all power supplies is required.  
MASTER POWER SUPPLY  
_
_
LOCAL  
_
MASTER  
+
+
+
M/S 1 M/S 2  
SLAVE  
CV  
CC  
SENSE  
OUT  
VREF  
A1 A2 A3 A4 A5  
+S  
-S  
CV  
CC  
REMOTE  
LOAD  
Figure 14 and Figure 15 show two and three supplies connected  
in auto-tracking with their negative output terminals connected  
together as a common or ground point. For two units in auto-  
tracking a fraction R2/(R1+R2) of the output of the master supply  
is provided as one of the inputs to the comparison amplifier of the  
slave supply, thus controlling the slave's output. The master sup-  
ply in an auto-tracking operation must be the positive supply hav-  
ing the largest output voltage. Turn-up and turn-down of the  
power supplies are controlled by the master supply. In order to  
maintain the temperature coefficient and stability specifications of  
the power supply, the external resistor should be stable, low  
noise, low temperature.  
R1 R2  
LOAD  
SLAVE POWER SUPPLY  
MASTER  
_
_
LOCAL  
_
+
+
+
M/S 1 M/S 2  
SLAVE  
CV  
CC  
SENSE  
+S  
OUT  
-S  
CV  
CC VREF A1 A2 A3 A4 A5  
REMOTE  
Figure 14. Auto-Tracking Operation of Two Supplies  
Determining Resistors. External resistors control the fraction of  
the master unit's voltage that is supplied from the slave unit. For  
two units in auto-tracking the ratio R1 and R2 is  
MASTER POWER SUPPLY  
_
_
_
LOCAL  
MASTER  
+
+
+
R2/(R1+R2 = (Vs/Vm)  
Where Vm = master output voltage  
Vs = slave output voltage  
M/S 1 M/S 2  
SLAVE  
CV  
CC  
REMOTE  
SENSE  
+S  
OUT  
-S  
CV  
CC VREF A1 A2 A3 A4 A5  
LOAD  
R1 R2  
LOAD  
SLAVE POWER SUPPLY(S1)  
_
_
_
LOCAL  
MASTER  
+
+
+
It is recommended to connect a 0.1 µF capacitor in paral-  
lel with R2 in two supplies operation or R2 and R4 in  
three supplies operation to ensure the stable operation.  
M/S 1 M/S 2  
SLAVE  
CV  
CC  
REMOTE  
SENSE  
Setting Voltage and Current. Use the master unit's VOLTAGE con-  
trol to set the output voltage from both units. When the master is in  
CV operation, the master's output voltage(Vm) is the same as its  
voltage setting, and the slave's output voltage for two units operation  
is Vm(R2/(R1+R2)). The VOLTAGE control of the slave unit is dis-  
abled. Set the CURRENT controls of master and slave units above  
the required currents to assure CV operation of master and slave  
units.  
+S  
OUT  
-S  
CV  
CC VREF A1 A2 A3 A4 A5  
R3 R4  
LOAD  
SLAVE POWER SUPPLY(S2)  
_
_
_
LOCAL  
MASTER  
+
+
+
M/S 1 M/S 2  
SLAVE  
CV  
CC  
REMOTE  
SENSE  
+S  
OUT  
-S  
CV  
CC VREF A1 A2 A3 A4 A5  
Overvoltage Protection. Set the OVP shutdown voltage in each  
unit so that it shuts down at a voltage higher than its output volt-  
age during auto-tracking operation. When a master unit shuts  
down, it programs any slave units to zero output. When a slave  
unit shuts down, it shuts down only itself.  
R2  
R1+R2  
R4  
Vm = masters unit's output voltage  
Vs1 = slave(S1) unit's output voltage  
Vs2 = slave(S2) unit's output voltage  
Vs1 =  
Vm  
Where  
Vs2 =  
Vs1  
R3+R4  
Figure 15. Auto-Tracking Operation of Three Supplies  
Remote Sensing. To include remote sensing with auto-tracking  
operation independently, set up each unit for remote sensing  
according to the remote-sensing instructions under previous  
paragraph.  
1-13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
a. The output impedance of the power supply decreases with  
increasing frequency.  
b. The recovery time of the output voltage is longer for load  
resistance changes.  
LOAD CONSIDERATIONS  
This section provides information on operating your supply with  
various types of loads connected to its output.  
c. A large surge current causing a high power dissipation in the  
load occurs when the load resistance is reduced rapidly.  
PULSE LOADING  
The power supply will automatically cross over from constant-  
voltage to constant current operation in response to an increase  
(over the preset limit) in the output current. Although the preset  
limit may be set higher than the average output current, high peak  
currents (as occur in pulse loading) may exceed the preset cur-  
rent limit and cause cross over to occur. If this cross over limiting  
is not desired, set the preset limit for the peak requirement and  
not the average.  
REVERSE VOLTAGE LOADING  
A diode is connected across the output terminals with reverse  
polarity. This diode protects the output electrolytic capacitors and  
the series regulator transistors from the effects of a reverse volt-  
age applied across the output terminals. For example, in series  
operation of two supplies, if the AC is removed from one supply,  
the diode prevents damage to the unenergized supply which  
would otherwise result from a reverse polarity voltage.  
REVERSE CURRENT LOADING  
An active load connected to the power supply may actually  
deliver a reverse current to the power supply during a portion of  
its operating cycle. An external source can not be allowed to  
pump current into the supply without loss of regulation and possi-  
ble damage to the output capacitor of the power supply. To avoid  
these effects, it is necessary to preload the supply with a dummy  
load resistor so that the power supply delivers current through the  
entire operating cycle of the load devices.  
Since series regulator transistors cannot withstand reverse volt-  
age, another diode is connected across the series transistor. This  
diode protects the series regulators in parallel or auto-parallel  
operation if one supply of the parallel combination is turned on  
before the other.  
BATTERY CHARGING  
The power supply's OVP circuit contains a crowbar SCR, which  
effectively shorts the output of the supply whenever the OVP trips. If  
an external voltage source such as a battery is connected across the  
output, and OVP inadvertently triggered, the SCR will continuously  
sink a large current from the source; possibly damaging the supply.  
To avoid this a diode must be connected in series with the output as  
shown in Figure 17.  
Figure 16. Reverse Current Loading Solution  
Figure 17. Recommended Protection Circuit for  
Battery Charging  
OUTPUT CAPACITANCE  
An internal capacitor, connected across the output terminals of  
the power supply, helps to supply high-current pulses of short  
duration during constant voltage operation. Any capacitance  
added externally will improve the pulse current capability, but will  
decrease the safety provided by the current limiting circuit. A  
high-current pulse may damage load components before the  
average output current is large enough to cause the current limit-  
ing circuit to operate.  
The effect of the output capacitor during constant current opera-  
tion are as follows:  
1-14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SERVICE INFORMATION  
Figure A-1. Block Diagram  
The main secondary winding of the power transformer has  
PRINCIPLES OF OPERATION  
(Block Diagram Overview)  
three sections (N1, N2, and N3), each of which has a different  
turns ratio with respect to the primary winding. At the begin-  
ning of each half-cycle of the input ac, the control circuit  
determines whether one pair, both or none of the SCR will be  
fired. If neither SCR is fired, the bridge diode (CR13) receives  
an ac input voltage that is determined by N1 turns and the  
input capacitors charge to a corresponding level. If SCR  
CR15 and CR18 are fired, input capacitors charge to the volt-  
age determined by N1+N2 turns. Similarly, if CR10 and CR12  
are fired the capacitors are charged by N1 + N3. Finally, if all  
SCRs are fired simultaneously, input capacitors charge to its  
highest voltage level determined by N1 + N2 + N3 turns.  
Throughout this discussion, refer to both the block diagram of  
Figure A-1 and the schematic diagrams at the rear of the  
manual. The input ac line voltage is first applied to the prereg-  
ulator which operates in conjunction with the SCR control cir-  
cuit (preregulator control circuit) to rectify the tap switched AC  
voltage. This preregulator minimizes the power dissipated in  
the series regulating elements by controlling the dc level  
across the input filter capacitor, depending on the output volt-  
age.  
To achieve this, tap switching is accomplished by four SCRs  
and one bridge diode (CR10, CR12, CR15, CR18 and CR13)  
and the SCR control circuit. By selecting different SCR firing  
combinations from SCR control circuit, these circuits allow the  
input capacitors (C7 and C8) to charge to one of four discrete  
voltage levels, depending on the output voltage required.  
The SCR control circuit determines which SCRs are to be  
fired by monitoring the output voltage and comparing these  
values against a set of three internally derived reference lev-  
els. These three reference levels are translated into boundary  
lines to allow the output characteristic to be mapped into four  
operating regions (Figure A-2). The boundary lines, which are  
invisible to the user, are divided into four operating regions  
(V1, V2, V3, and V4) to minimize the power dissipation in the  
A-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
series pass transistors. Whenever the output voltage is below  
the sloping V1 line, the control circuit inhibits four SCRs and  
the input capacitors charge to a voltage determined by N1.  
Figure A-2 indicates the windings that are connected as a  
result of the other voltage decisions.  
Full protection against any overload condition is inherent in  
the Constant Voltage/Constant Current design principle since  
there is not any load condition that can cause an output which  
lies outside the operating region. For either constant voltage  
or constant current operation, the proper choice of front panel  
voltage and current control settings insures optimum pro-  
tection for the load device as well as full protection for the  
power supply.  
The reference and bias circuit provides stable reference volt-  
ages which are used by the constant voltage/current error  
amplifier circuits for comparison purpose. The display circuit  
provides an indication of output voltage and current for con-  
stant voltage or constant current operating modes.  
An operator error or a component failure within the regulating  
feedback loop can drive a power supply's output voltage to  
many times its preset value. The overvoltage protection cir-  
cuit is to protect the load against this possibility. The circuit  
insures that the power supply voltage across the load will  
never exceed a preset limit.  
Figure A-2. Output Power Plot  
Diode CR19 is connected across the output terminals in  
reverse polarity. It protects the output electrolytic capacitor  
and the series regulator transistors from the effects of a  
reverse voltage applied across the output terminals.  
The series regulators (Q1 and Q4) are part of a feedback loop  
which consists of the driver and the Constant Voltage/Con-  
stant Current error amplifier. The series regulator feedback  
loop provides "fine and fast" regulation of the output while the  
preregulator feedback loop handles large, relatively slow, reg-  
ulation demands.  
The display power circuit provides voltage which is used by A/  
D converter and LED drive.  
The regulator is made to alter its conduction to maintain a  
constant output voltage or current. The voltage developed  
across the current sampling resistors (R58 and R59) is the  
input to the constant current error amplifier. The constant volt-  
age error amplifier obtains its input by sampling the output  
voltage of the supply.  
MAINTENANCE  
INTRODUCTION  
This section provides performance test and calibration proce-  
dures and troubleshooting information. The following opera-  
tion verification tests comprise a short procedure to verify that  
the power supply is performing properly, without testing all  
specified parameters.  
Any changes in output voltage or current are detected and  
amplified by the constant voltage or constant current error cir-  
cuit and applied to the series regulator in the correct phase  
and amplitude to counteract the change in output voltage or  
current.  
If a fault is detected in the power supply while making the  
performance check or during normal operation, proceed to  
the troubleshooting procedures. After troubleshooting, per-  
form any necessary adjustments and calibrations. Before  
returning the power supply to normal operation, repeat the  
performance check to ensure that the fault has been properly  
corrected and that no other faults exist.  
Two error amplifiers are included in a CV/CC supply, one for  
controlling output voltage, the other for controlling output cur-  
rent. Since the constant voltage amplifier tends to achieve  
zero output impedance and alters the output current when-  
ever the load resistance changes, while the constant current  
amplifier causes the output impedance to be infinite and  
changes the output voltage in response to any load resis-  
tance change, it is obvious that the two amplifiers can not  
operate simultaneously. For any given value of load resis-  
tance, the power supply must act either as a constant voltage  
source or as a constant current source - it can not be both;  
transfer between these two modes is accomplished at a value  
of load resistance equal to the ratio of the output voltage con-  
trol setting to the output current control setting.  
Test Equipment Required  
The following Table A-1 lists the equipment required to perform  
the tests and adjustments of this section. You can separately  
identify the equipment for performance tests, calibration, and  
troubleshooting in the USE column of the table.  
Operation Verification Tests  
The following tests assure that the power supply is per-  
forming properly. They do not, however, check all the speci-  
fied parameters tested in the complete performance test  
described below. Proceed as follows:  
A-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
a. Perform turn-on checkout procedure given in page 1-7.  
b. Perform the CV and CC Load Regulation perfor-  
mance tests given in the following paragraphs  
respectively.  
Electronic Load. The test and calibration procedures use an  
electronic load to test the supply quickly and accurately. An  
electronic load is considerably easier to use than load resis-  
tor. It eliminates the need for connecting resistors or rheostats  
in parallel to handle the power, it is much more stable than  
carbon-pile load, and it makes easy work of switching  
between load conditions as is required for the load regulation  
and load transient response tests.  
PERFORMANCE TESTS  
The following paragraphs provide test procedures for verify-  
ing the power supply's compliance with the specifications of  
Table 1. Please refer to adjustment and calibration or trouble-  
shooting procedure if you observe any out of specification  
performance.  
Current Monitoring Resistor Rs. To eliminate output-current  
measurement error caused by voltage drops in the leads and  
connections, connect the current monitoring (sampling) resis-  
tor between -OUT and the load as a four-terminal device. Fig-  
ure A-3 shows correct connections. Connect the current  
monitoring test leads inside the load lead connections directly  
at the monitoring resistor element. Select a resistor with sta-  
ble characteristics and lower temperature coefficient (see  
Table A-1).  
Measurement Techniques  
Setup for All Tests. Measure the output voltage directly at the  
+S and -S terminals on the rear panel; in this way the monitoring  
device sees the same performance as the feedback amplifier  
within the power supply. Failure to connect the monitoring device  
to the proper points shown in Figure A-3 will result in the mea-  
surement not of the power supply characteristics, but of the  
power supply plus the resistance of the leads between its output  
terminals and the point of connection.  
Use separate leads to all measuring devices to avoid the sub-  
tle mutual coupling effects that may occur between measur-  
ing devices unless all are returned to the low impedance  
terminals of the power supply. Twisted pairs or shielded cable  
should be used to avoid pickup on the measuring leads.  
Figure A-3. Current Monitoring Resistor Connections  
Table A-1. Test Equipment Required  
TYPE  
Oscilloscope  
REQUIRED CHARACTERISTICS  
Sensitivity : 1 mV  
USE  
P, T  
RECOMMENDED MODEL  
Agilent 54600A  
Bandwidth : 20 MHz/100 MHz  
Input : Differential, 50 ohm and 100 ohm  
RMS Voltmeter  
True rms, 20 MHz bandwidth  
Sensitivity : 1 mV  
P
Accuracy : 5%  
Multimeter  
Resolution : 100 nV  
Accuracy : 0.0035%  
P, A, T  
P
Agilent 34401A  
Agilent 6063A  
Electronic Load  
Voltage Range : 240 Vdc  
Current Range : 10 Adc  
Open and short switches  
Transient on/off  
Load Resistor (R )  
1.3 ohm 60 W, 6.6 ohm 60 W, 20.5 ohm 60 W,  
60 ohm 60 W  
P
L
Current Monitoring  
0.1 ohm 0.1% 10 W, 1 ohm 1% 5 W  
P, A  
(Sampling) Resistor (R )  
S
Variable Voltage  
Auto Transformer  
Range : 85-130 and 200-260 Volts  
P, T  
* P = Performance testing A = Calibration adjustments T = Troubleshooting.  
A-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Line Regulation (Source Effect)  
CONSTANT VOLTAGE (CV) TESTS  
Definition: Line regulation is the change in the steady state  
value of dc output voltage due to a change in ac input voltage  
from a minimum to a maximum value(±10% of nominal volt-  
age).  
CV Setup. For all CV tests set the output current at full rated  
output to assure CV operation. The onset of constant current  
can cause a drop in output voltage, increased ripple, and  
other performance changes not properly ascribed to the con-  
stant voltage operation of the supply.  
Test Parameter:  
Measured Variable: Output Voltage  
Expected Results: Less than 0.01% plus 2 mV  
Test Procedure:  
Load Regulation (Load Effect)  
Definition: CV Load regulation is the change in the steady  
state value of dc output voltage due to a change in load resis-  
tance from open circuit to full load or from full load to open cir-  
cuit.  
a. Connect the test equipment as shown in Figure A-4.  
Operate the electronic load in constant current mode  
and set its current to the full rated value of the power  
supply.  
Test Parameters:  
b. Connect the supply to the ac power line through a  
variable autotransformer which is set for low line volt-  
age(104 Vac for nominal 115 Vac, 90 Vac for nominal  
100 Vac, and 207 Vac for nominal 230 Vac).  
c. Turn the supply's power on and turn CURRENT con-  
trol fully clockwise.  
d. Adjust VOLTAGE control until the front panel VOLTS  
display indicates exactly the maximum rated output  
voltage.  
e. Record voltage indicated on the digital voltmeter.  
f. Adjust autotransformer to high line voltage(127 Vac  
for nominal 115 Vac, 110 Vac for nominal 100 Vac,  
and 253 Vac for nominal 230 Vac).  
g. When the reading settles, record the output voltage  
again. Check that the two recorded readings differ  
less than 0.01% of output voltage plus 2 mV.  
Measured Variable: Output Voltage  
Expected Results: Less than 0.01% plus 2 mV  
Test Procedure:  
a. Connect the test equipment as shown in Figure A-4.  
Operate the electronic load in constant current mode  
and set its current to the full rated value of the power  
supply (6 A for E3614A, 3 A for E3615A, 1.7 A for  
E3616A and 1 A for E3617A).  
b. Turn the supply's power on and turn CURRENT con-  
trol fully clockwise.  
c. Turn up output voltage to the full rated value (8 V for  
E3614A, 20 V for E3615A, 35 V for E3616A and 60 V  
for E3617A) as read on the digital voltmeter.  
d. Record the output voltage at the digital voltmeter.  
e. Operate the electronic load in open(input off) mode.  
f. When the reading settles, record the output voltage on  
the digital voltmeter again. Check that the two recorded  
readings differ less than 0.01% of output voltage plus 2  
mV.  
Load Transient Response Time  
Definition : This is the time for the output voltage to return to  
within a specified band around its voltage following a change  
from full load to half load or half load to full load.  
Test Parameter:  
POWER SUPPLY  
LOCAL  
MASTER  
Measured Variable: Output Voltage Transients  
Expected Results: Less than 50 usec (at 15 mV from  
base line)  
UNDER TEST  
Test Procedure:  
a. Connect the test equipment as shown in Figure A-4,  
but replace the DVM with the oscilloscope. Operate  
the electronic load in constant current mode.  
b. Turn the supply's power on and turn CURRENT con-  
trol fully clockwise.  
SENSE  
CV  
CC  
M/S 1  
M/S 2  
REMOTE  
SLAVE  
+
-
+
-
+
-
c. Turn up output voltage to the full rated value.  
d. Set the electronic load to transient operation mode  
between one half of supply's full rated value and sup-  
ply's full rated value at a 1 KHz rate with 50% duty  
cycle.  
e. Set the oscilloscope for ac coupling, internal sync and  
lock on either the positive or negative load transient.  
f. Adjust the oscilloscope to display transients as in Fig-  
ure A-5.  
+S  
-S  
CC  
VREF A1 A2 A3 A4 A5  
OUT  
CV  
+
DIGITAL  
-
VOLTMETER  
-
TO  
Rs  
DVM  
+
Model  
Rs  
-
+
g. Check that the pulse width of the transients at 15 mV  
from the base line is no more than 50 usec as shown.  
E3614A, 15A, 16A  
E3617A  
0.1 ohm 0.1% 10W  
1 ohm 1% 5W  
ELECTRONIC  
LOAD  
Figure A-4. Basic Test Setup  
A-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Figure A-5. Load Transient Response Time Waveform  
PARD(Ripple and Noise)  
Definition: PARD is the Periodic and Random Deviation of  
the dc output voltage from its average value, over a specified  
bandwidth and with all other parameters maintained constant.  
Constant voltage PARD is measured in the root-mean-  
square(rms) or peak-to-peak(pp) values over a 20 Hz to 20  
MHz bandwidth. Fluctuations below the lower frequency limit  
are treated as drift.  
PARD(RMS) Measurement  
The rms measurement is not an ideal representation of the  
noise, since fairly high output noise spikes of short duration  
could be present in the ripple and not appreciably increase  
the rms value.  
Test Parameter:  
Measured Variable: Output Voltage(rms)  
Expected Results: Less than 200 µV rms  
Test Procedure:  
Figure A-6. CV PARD RMS Measurement Test Setup  
a. Connect the test equipment as shown in Figure A-6.  
b. Turn the supply's power on and turn CURRENT con-  
trol fully clockwise.  
c. Turn up output voltage to the full rated value. Check  
that the supply's CV indicator remains lighted.  
Reduce VOLTAGE control if not lighted.  
d. Check that the rms noise voltage at the true rms volt-  
meter is less than 200BµV.  
PARD(Peak-to-Peak) Measurement  
The peak-to-peak measurement is particularly important for  
applications where noise spikes could be detrimental to a  
sensitive load, such as logic circuitry.  
Test Parameter:  
Measured Variable: Output voltage(peak-to-peak)  
Expected Results: Less than 1 mV p-p (20 Hz-20 MHz)  
Test Procedure:  
A-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
a. Connect the test equipment as shown in Figure A-7.  
b. Turn the supply's power on and turn CURRENT con-  
trol fully clockwise.  
c. Turn up output voltage to the full rated value. Check  
that the supply's CV indicator remains lighted.  
Reduce VOLTAGE control if not lighted.  
d. Set the oscilloscope to AC mode and bandwidth to 20  
MHz.  
CONSTANT CURRENT (CC) TESTS  
CC Setup. Constant current tests are analogous to constant  
voltage tests, with the supply's output short circuited and the  
voltage set to full output to assure CC operation. For output  
current measurements the current monitoring resistor must  
be treated as a four terminal device. Refer to the "Measure-  
ment Techniques" for details. All constant current measure-  
ments are made in terms of the change in voltage across this  
resistor; the current performance is calculated by dividing  
these voltage changes by ohmic value of Rs.  
e. Check that the peak-to-peak noise is less than 1 mV.  
Load Regulation (Load Effect)  
Definition: CC Load regulation is the change in the steady state  
value of dc output current due to a change in load resistance  
from short circuit to full load or from full load to short circuit.  
Test Parameter:  
Measured Variable: Output Current  
Expected Results: Less than 0.01% plus 250 µA  
Test Procedure:  
a. Connect the DVM across Rs in Figure A-4. Operate  
the electronic load in constant voltage mode and set  
its voltage to the full rated value of power supply.  
b. Turn the supply's power on and turn VOLTAGE con-  
trol fully clockwise.  
c. Turn up output current to the full rated value. Check  
that the AMPS display reads full rated values and CC  
indicator remains lighted. Reduce CURRENT control  
if not lighted.  
d. Record the voltage across Rs and convert it to cur-  
rent by dividing this voltage by Rs.  
e. Operate the electronic load in short (input short)  
mode.  
Figure A-7.BCV PARD Peak-to-Peak Measurement Test  
Setup  
f. When the reading settles, record voltage across Rs  
again and convert it current. Check that the two  
recorded readings differ less than 0.01% of output  
current plus 250 µA.  
CV Drift (Stability)  
Definition: The change in output voltage (dc to 20 Hz) for the  
first 8 hours following a 30-minute warm-up period with con-  
stant input line voltage, constant load resistance and constant  
ambient temperature.  
Line Regulation (Source Effect)  
Definition: Line regulation is the change in the steady state  
value of dc output current due to a change in ac input voltage  
from the minimum to maximum value(±10% of nominal voltage).  
Test Parameter:  
Measured Variable: Output Voltage  
Expected Results: Less than 0.1% plus 5 mV  
Test Procedure:  
Test Parameter:  
Measured Variable: Output Current  
Expected Results: Less than 0.01% plus 250 µA  
Test Procedure:  
a. Connect the DVM across Rs in Figure A-4.  
b. Operate the electronic load in constant current mode  
and set its current to the full rated value of power sup-  
ply.  
c. Turn the supply's power on and turn CURRENT con-  
trol fully clockwise.  
d. Turn up output voltage to the full rated value as read  
on the digital voltmeter.  
e. After a 30-minute warm-up, note the voltage on DVM.  
f. The output voltage reading should deviate less than  
0.1% plus 5 mV from the reading obtained in step e  
over a period of 8 hours.  
a. Connect the DVM across Rs in Figure A-4. Operate  
the electronic load in constant voltage mode and set  
its voltage to the full rated value of power supply.  
b. Connect the supply to the ac power line through a  
variable autotransformer that set for low line volt-  
age(104 Vac for nominal 115 Vac, 90 Vac for nominal  
100 Vac, and 207 Vac for nominal 230 Vac).  
c. Turn the supply's power on and turn VOLTAGE con-  
trol fully clockwise.  
d. Turn up output current to the full rated value. Check  
that the AMPS display reads full rated values and CC  
indicator remains lighted. Reduce CURRENT control  
if not lighted.  
A-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
e. Record output voltage across Rs and convert it to  
current by dividing this voltage by Rs.  
f. Adjust autotransformer to the high line voltage (127  
Vac for nominal 115 Vac, 110 Vac for nominal 100  
Vac, and 253 Vac for nominal 230 Vac).  
CC Drift (Stability)  
Definition: The change in output current for the first 8 hours fol-  
lowing a 30-minute warm-up with constant input line voltage,  
constant load resistance and constant ambient temperature.  
g. When the reading settles, record the voltage across  
Rs again and convert it current. Check that the two  
recorded readings differ less than 0.01% of output  
current plus 250 µA.  
Test Parameter:  
Measured Variable: Output Current  
Expected Results: Less than 0.1% plus 10 mA  
Test Procedure:  
a. Connect the DVM across Rs in Figure A-4. Operate  
the electronic load in constant voltage mode and set  
its voltage to the full rated value of the power supply.  
b. Turn the supply's power on and turn VOLTAGE con-  
trol fully clockwise.  
c. Turn up output current to the full rated value.  
d. After a 30-minute warm-up, note the voltage on DVM  
and convert it to current by dividing this voltage by Rs.  
e. The converted output current should deviate less than  
0.1% plus 10 mA from the current obtained in step d  
over a period of 8 hours.  
PARD(Ripple and Noise)  
Definition : The residual ac current which is superimposed  
on the dc output current of a power supply. Constant current  
PARD is specified as the root-mean-square(rms) output cur-  
rent in a frequency range of 20 Hz to 20 MHz with the supply  
in CC operation.  
PARD(RMS) Measurement  
Test Parameter:  
Measured Variable: Output Current(rms)  
Expected Results: E3614A: Less than 5 mA rms  
E3615A: Less than 2 mA rms  
ADJUSTMENT AND CALIBRATION  
PROCEDURE  
Adjustment and calibration may be required after perfor-  
mance testing, troubleshooting, or repair and replacement.  
Perform those adjustments that affect the operation of the  
faulty circuit and no others. To remove the top cover, refer to  
"Line Voltage Option Conversion" paragraph.  
E3616A: Less than 500 µA rms  
E3617A: Less than 500 µA rms  
Test Procedure:  
a. Connect the test equipment as shown in Figure A-8.  
b. Turn the supply's power on and turn VOLTAGE con-  
trol fully clockwise.  
c. Turn up output current to the full rated value. Check  
that the CC indicator remains lighted. Reduce CUR-  
RENT control if not lighted.  
d. Record rms voltage across Rs and convert it to cur-  
rent by dividing this voltage by Rs.  
e. Check that the rms noise current is less than 5 mA  
rms for E3614A, 2 mA rms for E3615A and 500 µA  
rms for E3616A and E3617A respectively.  
Maintenance described herein is performed with  
power supplied to the supply, and protective covers  
removed. Such maintenance should be performed  
only service-trained personnel who are aware of the  
hazards involved (for example, fire and electrical  
shock). Where maintenance can be performed with-  
out power applied, the power should be removed.  
Figure A-8. CC PARD RMS Measurement Test Setup  
Figure A-9. Calibration Test Setup  
A-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Overall Troubleshooting Procedure  
Ammeter and CC Set Calibration  
To calibrate ammeter and CC set, proceed as follows:  
a. Connect test setup on Figure A-9.  
To locate the cause of trouble follow steps 1, 2, and 3 in  
sequence. Before attempting overall troubleshooting, ensure  
that the rear-panel switches M/S 1 and M/S 2 be set to MAS-  
TER position and CV, CC, and SENSE to LOCAL position.  
1. Check that input power is available, and check the  
power cord and rear panel line fuse. When replacing  
line fuse, be certain to select fuse of proper rating for  
line voltage being used.  
b. Turn VOLTAGE and CURRENT control fully clock-  
wise.  
c. Turn on the supply and to calibrate ammeter adjust  
R5 on the display board until front panel AMPS dis-  
play reads exactly DVM value divided by Rs.  
d. To calibrate CC Set adjust R69 on the main board  
until front panel AMPS display reads exactly DVM  
value divided by Rs while depressing OVP/CC Set  
switch.  
2. In almost all cases, the trouble source can be caused  
by the dc bias or reference voltages; thus, it is a good  
practice to check voltages in Table A-2 before pro-  
ceeding with step 3.  
3. Disconnect the load and examine Table A-3 to deter-  
mine your symptom, then check the probable cause.  
Voltmeter and OVP Set Calibration  
To calibrate voltmeter and OVP set, proceed as follows:  
a. Disconnect Rs from test setup on Figure A-9 and  
connect DVM across output terminal of the supply.  
b. Turn on the supply.  
Reference and Bias Circuit  
a. Make an ohmmeter check to be certain that neither  
the positive and negative output terminal is grounded.  
b. Turn front panel VOLTAGE and CURRENT controls  
fully clockwise.  
c. To calibrate voltmeter for E3614A, adjust R16 on the  
display board until front panel VOLTS display reads  
exactly DVM value. To calibrate voltmeter for  
c. Turn on power supply (no load connected).  
d. Proceed as instructed in Table A-2.  
E3615A, E3616A and E3617A set the output voltage  
below 18V (ex, 15V), and adjust R16 on the display  
board until front panel VOLTS display reads exactly  
DVM value. Next, set the output voltage above 20V  
(ex, 21V) and adjust R17 on the display board until  
front panel VOLTS display reads exactly DVM value.  
d. To calibrate OVP Set, turn down the OVP Adjust  
screwdriver control on the front panel slowly until the  
OVP circuit trips. Record the output voltage when the  
OVP trip occurs. Then adjust R97 on the main board  
until front panel VOLTS display reads exactly OVP  
trip voltage while depressing OVP/CC Set switch.  
Regulating Loop Troubles  
If the voltages in Table A-2 have been checked to eliminate  
the reference and bias circuits as a source of trouble; the mal-  
function is caused by either the series regulator or preregula-  
tor feedback loop. Because the interaction between these two  
loops makes logical troubleshooting difficult, the following  
steps help you to locate the source of troubles in these two  
feedback loops. Once the trouble has been located to one of  
the feedback loops, the operation of either loop can be ana-  
lyzed independently. This method should be followed when-  
ever a low output voltage condition exists. Notice that  
troubleshooting can proceed directly as described in Table A-  
4 whenever a high output voltage condition exists.  
1. Turn on the power supply with full load connected  
and increase output voltage by turning up the front  
panel voltage control. The output voltage is clamped  
and CV indicator is turned off at some output voltage  
(below full rated output voltage). If this is the case,  
the series regulator feedback loop is operating  
TROUBLESHOOTING  
Before attempting to troubleshoot the power supply, ensure  
that the fault is with the supply and not with an associated cir-  
cuit. The performance test enables this to be determined  
without having to remove the covers from the supply.  
normally and the trouble condition is probably due to  
a defect in the preregulator feedback loop (refer to  
Table A-6). If the output voltage remains in low stage,  
and varying the front panel voltage control has little or  
no effect, then the trouble is probably in the series  
regulator feedback loop. Refer to Table A-5.  
The applicable test points are identified by encircled  
numbers on the schematic diagrams at the rear of the  
manual, Figure A-10, Figure A-11, Figure A-12, and  
Figure 13.  
2. Measure the voltage between TP2 and TP1 (shown  
on the schematic diagram at the rear of the manual)  
with full load with oscilloscope while increasing the  
output voltage from 0 to full rated voltage. The volt-  
age measured has step changes three times during 0  
to full output voltage swing. If this is the case, prereg-  
ulator feedback loop is operating normally. If this is  
not the case, the trouble is probably in the preregula-  
tor feedback loop. Refer to Table A-6.  
A good understanding of the principles of operation is a help-  
ful aid in troubleshooting, and it is recommended that princi-  
ples of operation in this manual be reviewed before  
attempting to troubleshoot the supply. Once the principles of  
operation are understood, refer to the overall troubleshooting  
procedures paragraph to locate the symptom and probable  
cause.  
Once the defective component has been located (by means  
of visual inspection or trouble analysis) replace it and recon-  
duct the performance test. After a component is replaced,  
perform the meter calibration.  
A-8  
Download from Www.Somanuals.com. All Manuals Search And Download.  
After the trouble has been isolated to one of the feedback  
loops, troubleshooting can proceed as described in Tables A-  
4, A-5, or A-6.  
series regulator backwards a stage at a time, since loop fail-  
ures occur more often at the higher power levels.  
Preregulator Feedback Loop. The preregulator feedback  
loop (SCR control circuit) can be conveniently checked using  
Table A-6. As indicated in Table A-6, the control circuit is  
checked by starting with the waveform at point 7 and point 6  
(shown on the schematic diagram) and tracing forwards and  
backwards from this point.  
Series Regulating Feedback Loop. When troubleshooting  
the series regulating loop, it is useful to open the loop since  
measurements made anywhere within a closed loop may  
appear abnormal. With a loop closed, it is very difficult to sep-  
arate cause from effect. As described in Tables A-4 and A-5,  
the conduction or cutoff capability of each stage is checked  
by shorting or opening a previous stage, as follows:  
1. Shorting the emitter to collector of a transistor simu-  
lates saturation, or the full ON condition.  
Overvoltage Protection Circuit Troubles  
When troubleshooting the overvoltage protection circuit, it is  
useful to check the turn-on overshoot control circuit which  
includes U20 and Q10. The function of the control circuit is to  
slow down the rising speed of the +15 V bias the moment the  
power is turned on. This function prevents the supply from  
false OVP tripping the moment the power is turned on. After  
the troubles has been isolated to overvoltage protection cir-  
cuit, troubleshooting can proceed as described in Table A-7.  
2. Shorting the emitter to base of a transistor cuts it off,  
and simulates an open circuit between emitter and  
collector.  
Although a logical first choice might be to break the loop  
somewhere near its mid-point, and then perform successive  
subdividing tests, it is more useful to trace the loop from the  
Table A-2. Reference and Bias Circuit Troubleshooting  
METER  
COMMON  
METER  
POSITIVE  
NORMAL INDICATION  
NORMAL RIPPLE  
(p-p)  
PROBABLE CAUSE  
TP6  
TP6  
TP6  
TP6  
TP6  
point 2  
+15.0 +/- 0.3 Vdc  
-12.0 +/- 0.3 Vdc  
+10.5 +/- 0.2 Vdc  
-5.1 +/- 0.5 Vdc  
+5.0 +/- 0.3 Vdc  
2 mV  
2 mV  
2 mV  
2 mV  
4 mV  
Check U13, CR31, and CR32.  
Check +15 V bias or U14.  
Check +15 V bias, U11, and U14.  
Check -12 V bias or VR1.  
Check U1 and CR2.  
point 4  
TP7  
point 3  
point 5  
Table A-3. Overall Troubleshooting  
CHECKS AND PROBABLE CAUSES  
SYMPTOM  
High Output Voltage  
a. Check series regulator feedback loop or preregulator feedback loop.  
b. Refer to "Regulating Loop Troubles" paragraph or Table A-4 or A-6 as instructed.  
Low and No Output Voltage  
High Ripple  
a. If output is zero, check fuse.  
b. Check series regulator feedback loop or preregulator loop.  
Refer to "Regulating Loop Troubles" paragraph or Table A-5 or A-6 as instructed.  
c. Check CR20 shorted.  
a. Check operating setup for ground loops.  
b. If output floating, connect 1 µF capacitor between output and ground.  
c. Ensure that the supply is not crossing over to constant current mode  
under loaded conditions.  
d. Check for low voltage across C7 or Q1 and Q4.  
e. Check for excessive ripple on reference voltages (Table A-2).  
Poor Line Regulation  
(Constant Voltage)  
a. Check +10 V reference voltage.  
b. Check U9.  
A-9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table A-3. Overall Troubleshooting (Cont’d)  
CHECKS AND PROBABLE CAUSES  
SYMPTOM  
Poor Load Regulation  
(Constant Voltage)  
a. Refer to "Measurements Techniques" paragraph.  
b. Check +10 V reference voltage.  
c. Ensure that the supply is not going into current limit.  
Poor Load Regulation  
(Constant Current)  
a. Check +10 V reference voltage.  
b. CR1, CR19, CR20, C2, C31 leaky.  
c. Ensure that the supply is not crossing over to constant voltage operation.  
Oscillates (Constant Voltage/  
Constant Current)  
a. Check C29 and C36 in constant voltage circuit.  
b. Check C31 and C33 in constant current circuit.  
Poor Stability  
(Constant Voltage)  
a. Check +10 V reference voltage.  
b. CR27, CR28, CR23, and CR26 leaky.  
c. U9 defective.  
d. Noisy programming resistor R83.  
Poor Stability  
(Constant Current)  
a. Check +10 V reference voltage.  
b. CR24, CR25, CR29, and CR30 leaky.  
c. U9 and U10 defective.  
d. Noisy programming resistor R85.  
Excessive heat  
OVP Shutdown  
a. Check preregulator control circuit. Refer to Table A-6.  
b. CR10, CR12, CR15, and CR18 short  
a. Check that the front panel OVP Adjust screw control is rotated fully clockwise.  
b. Check the overvoltage protection circuit.  
Refer to "Overvoltage Protection Circuit Troubles" paragraph or Table A-7.  
Table A-4. High Output Voltage Troubleshooting  
STEP  
ACTION  
RESPONSE  
PROBABLE CAUSE  
a. Q1 or Q4 shorted.  
1
Check turn off of Q1 and  
Q4 by shorting Q9 emitter  
to collector.  
a. Output voltage remains high.  
b. Output voltage decreases.  
b. Remove short and proceed to step 2.  
2
3
Check turn on of Q9 by  
shorting point 1 to -12 V.  
a. Output voltage remains high.  
b. Output voltage decreases.  
a. Q9 open.  
b. Remove short and proceed to step 3.  
Check voltage from pin 5  
to pin 6 of U9.  
a. Input voltage is positive.  
b. Input voltage is negative.  
a. U9B is defective.  
b. Turn down the voltage control fully  
counter clockwise. Check the voltage  
of U9 pin 1 is 0.  
Table A-5. Low Output Voltage Troubleshooting  
RESPONSE  
STEP  
ACTION  
PROBABLE CAUSE  
1
Check turn on of Q1 and  
Q4 by disconnecting emitter  
of Q9.  
a. Output voltage remains low.  
b. Output voltage increases.  
a. Q1 or Q4 open.  
b. Reconnect emitter lead and proceed to step 2.  
2
3
Check turn off of Q9 by  
shorting point 1 to +15 V.  
a. Output voltage remains low.  
b. Output voltage increases.  
a. Q9 shorted.  
b. Remove short and proceed to step 3.  
Eliminate constant current  
comparator as a source of  
trouble by disconnecting  
anode of CR22.  
a. Output voltage is increases.  
b. Output voltage remains low.  
a. Proceed to step 4.  
b. Reconnect lead and proceed to step 5.  
A-10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table A-5. Low Output Voltage Troubleshooting (Cont’d)  
STEP  
ACTION  
RESPONSE  
PROBABLE CAUSE  
4
Check voltage from pin 13 to  
pin 12 of U9.  
a. Measured voltage is positive.  
b. Measured voltage is negative.  
a. Check U9A is defective.  
b. Check U10 and U9D is defective.  
Check R85 is open.  
5
Check voltage from pin 6  
to pin 5 of U9.  
a. Measured voltage is positive.  
b. Measured voltage is negative.  
a. U9B is defective.  
b. Check U9C is defective.  
Table A-6. Preregulator/Control Circuit Troubleshooting  
MEASURE RESPONSE  
STEP  
PROBABLE CAUSE  
1
Set output voltage to 4.5 V +- 0.5 V for E3614A.  
Set output voltage to 10 V +- 1 V for E3615A.  
Set output voltage to 15 V +- 1 V for E3616A.  
Set output voltage to 26 V +- 5 V for E3617A.  
2
E3614A  
E3615A  
Waveform form from  
TP6(common) to point 6  
a. Normal firing pulse  
a. Check CR18, CR15, Q7, Q8  
for defective.  
b. No firing pulse  
b. Proceed to step 3.  
E3616A  
Voltage from TP6  
(common) to point 6  
a. High voltage (+0.7 V)  
a. CR15, CR18, U2, U21  
defective  
b. Low voltage (0 V)  
b. Proceed to step 3.  
3
4
5
6
Voltage from TP6(common) to  
U4 pin 1  
a. Low voltage (-12 V)  
b. High voltage (+5 V)  
a. U3 defective  
b. Proceed to step 4.  
Voltage from TP6(common) to  
U5 pin 1  
a. High voltage (+15 V)  
b. Low voltage (-12 V)  
a. U4 defective  
b. Proceed to step 5.  
Voltage from pin 6 to  
pin 7 of U5  
a. Measured voltage is positive.  
b. Measured voltage is negative.  
a. U5 defective  
b. U6 defective  
Set output voltage to 7 V +- 1 V for E3614A.  
Set output voltage to 16 V +- 2 V for E3615A.  
Set output voltage to 25 V +- 2 V for E3616A.  
Set output voltage to 44 V +- 5 V for E3617A.  
7
Waveform form from TP6  
(common) to point 7  
a. Normal firing pulse  
a. CR10, CR12, Q5, Q6  
defective  
b. No firing pulse  
b. Proceed to step 8.  
8
9
Voltage from TP6(common)  
to U4 pin 14  
a. Low voltage (-12 V)  
b. High voltage (+5 V)  
a. U3 defective  
b. Proceed to step 9.  
Voltage from TP6(common)  
to U5 pin 14  
a. High voltage (+15 V)  
b. Low voltage (-12 V)  
a. U4 defective  
b. Proceed to step 10.  
10  
Voltage from pin 8 to  
pin 9 of U5  
a. Measured voltage is positive.  
b. Measured voltage is negative.  
a. U5 defective  
b. U6 defective  
A-11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table A-7. Overvoltage Protection Circuit Troubleshooting  
STEP  
ACTION  
RESPONSE  
PROBABLE CAUSE  
1
Short U19 pin 4 to TP6.  
a. Shutdown release  
(OVP indicator OFF)  
b. Output voltage remains  
shutdown(0 V)  
a. U20 defective or C57  
shorted.  
b. Proceed to step 2.  
2
3
Measure the voltage from  
TP6(common) to TP9.  
a. High voltage (+5 V)  
b. Low voltage (0 V)  
a. U19 defective or proceed step 3.  
b. U4D defective.  
Measure the voltage from  
TP6(common) to TP8.  
a. Below +2.6 V  
b. Above +2.6 V  
a. U12 or U8 defective  
b. U18 defective  
Table A-8. Reference Designators  
REPLACEABLE PARTS  
A
C
Assembly  
Capacitor  
Diode  
INTRODUCTION  
This section contains information for ordering replacement  
parts. Table A-10 lists parts by reference designators and pro-  
vides the following information:  
CR  
DS  
F
Signaling Device(light)  
Fuse  
G
Pulse Generator  
Jack  
a. Reference designators. Refer to Table A-8.  
b. Agilent Technologies Part Number.  
c. Total quantity used in that assembly.  
d. Description.  
J
L
Inductor  
Q
Transistor  
Resistor  
R
e. Manufacturer's supply code number. Refer to Table  
A-9 for manufacturer's name and address.  
f. Manufacturer's part number or type.  
S
Switch  
T
Transformer  
Test Point  
Zener Diode  
Integrated Circuit  
Wire  
TP  
VR  
U
Mechanical and miscellaneous parts are not identified by ref-  
erence designator.  
W
ORDERING INFORMATION  
To order a replacement part, address order or inquiry to your local Agilent Technologies sales office (see lists at rear of this manual for  
addresses). Specify the following information for each part: Model, complete serial number of the power supply; Agilent Technologies  
part number; circuit reference designator; and description.  
Table A-9. Code List of Manufacturers  
CODE  
MANUFACTURER  
Texas Instruments Inc, Semicon Comp Div.  
ADDRESS  
Dallas, TX  
01295  
14936  
27014  
28480  
04713  
32997  
34371  
General Instruments Corp, Semicon Prod  
National Semiconductor Corporation  
Agilent Technologies  
Hicksville, N.Y.  
Santa Clara, CA  
Palo Alto, CA  
Phoenix, AZ  
Motorola Semiconductor Products  
Bourns Inc.  
Riverside, CA  
Melbourne, FL  
Harris Corp.  
A-12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table A-10. Replaceable Parts List  
Reference Designator  
Agilent Part  
Number  
Q'ty  
Description  
Model  
Mfr. P/N  
Mfr.  
Code  
E3614A  
1
1
1
1
60WBENCHPOWERSUPPLY-E3614AMODEL  
60WBENCHPOWERSUPPLY-E3615AMODEL  
60WBENCHPOWERSUPPLY-E3616AMODEL  
60WBENCHPOWERSUPPLY-E3617AMODEL  
E3615A  
E3616A  
E3617A  
E3614-60005 1  
E3615-60005 1  
E3616-60005 1  
E3617-60005 1  
MAIN BODY ASSY  
MAIN BODY ASSY  
MAIN BODY ASSY  
MAIN BODY ASSY  
KEYCAP-WHT  
KEYCAP-GRAY  
14  
28480  
28480  
28480  
28480  
28480  
28480  
15  
16  
17  
0371-3806  
0371-8624  
1
1
ALL  
ALL  
E361X-60003 1  
E361X-60009 1  
FRONT PANEL ASSY  
ALL  
ALL  
ALL  
ALL  
ALL  
ALL  
ALL  
ALL  
14  
28480  
28480  
DISPLAYBOARDASSY  
R84,85  
2100-4503  
2
RES-VAR10K5%10-TURNWW  
BINDINGPOSTRED  
3590S-A17-103 32997  
28480  
E3631-20011 1  
E3631-20012 1  
E3631-20013 1  
BINDINGPOSTBLACK  
28480  
BINDINGPOSTGREEN  
28480  
2950-0144  
5041-8621  
3
2
NUT-BINDINGPOST  
28480  
KNOB  
28480  
E3614-60002 1  
E3615-60002 1  
E3617-60002 1  
E3617-60002 1  
MAIN BOARD ASSY  
28480  
MAIN BOARD ASSY  
15  
28480  
MAIN BOARD ASSY  
16  
28480  
MAIN BOARD ASSY  
17  
28480  
5021-8128  
5021-8139  
0160-4835  
0180-4360  
0180-4355  
0180-4452  
0180-3595  
0160-0269  
0180-3990  
0180-4567  
0180-4568  
0180-4607  
0180-4566  
0180-3970  
0160-7449  
0160-4832  
0160-4835  
1
1
3
1
1
1
1
2
1
2
2
2
2
5
3
5
1
PCBMAINFORE3614A,E3615A  
PCBMAINFORE3616A,E3617A  
CAP-FXD.1UF+-10%50VCERX7R  
CAP-FXD1000UF25V+-20%AL-ELECTLT  
CAP-FXD470UF50V+-20%AL-ELECTLT  
CAP-FXD330UF63V+-20%AL-ELECTLT  
CAP-FXD220UF100V+-20%AL-ELECTLT  
CAP-FXD0.1UF+-20%500VCERZ5U  
CAP-FXD4700UF+-20%25VAL-ELECTLT  
CAP-FXD39000UF25V+-20%AL-ELECTLT  
CAP-FXD12000UF63V+-20%AL-ELECTLT  
CAP-FXD5600UF100V+-20%AL-ELECTLT  
CAP-FXD2700UF160V+-20%AL-ELECTLT  
CAP-FXD1UF+-20%50VAL-ELECTLT  
CAP-FXD0.33UF+-10%50VPOLYE-FL  
CAP-FXD0.01UF+-10%100VCERX7R  
CAP-FXD.1UF+-10%50VCERX7R  
CAP-FXD0.01UF+-10%100VCERX7R  
14,15  
16,17  
ALL  
14  
28480  
28480  
C1,3,47  
28480  
C2  
28480  
C2  
15  
28480  
C2  
16  
28480  
C2  
17  
28480  
C4,5  
ALL  
ALL  
14  
28480  
C6  
28480  
C7,8  
28480  
C7,8  
15  
28480  
C7,8  
16  
28480  
C7,8  
17  
28480  
C9,41,42,43,55  
C10,48,49  
C11,16,17,25,26  
C12  
ALL  
ALL  
14,15  
14,15  
ALL  
28480  
28480  
28480  
28480  
C13,14,15,30,32,3 0160-4832  
4,38,39,50,54  
10  
28480  
A-13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table A-10. Replaceable Parts List (Cont'd)  
Reference Designator  
Agilent Part  
Number  
Q'ty  
Description  
Model  
Mfr. P/N  
Mfr.  
Code  
C18,21,24,27  
C19,22  
C20,23  
C28  
0160-7077  
0160-4822  
0180-3970  
0160-6225  
0160-4832  
0160-7673  
0160-7075  
4
2
2
1
1
1
1
6
CAP-FXD.1UF+-10%630VPOLYE-FL  
CAP-FXD1000PF+-5%100VCERCOG  
CAP-FXD1UF+-20%50VAL-ELECTLT  
CAP-FXD0.33UF+-10%250VPOLYE-MET  
CAP-FXD0.01UF+-10%100VCERX7R  
CAP-FXD.047UF+-10%100VPOLYP-MET  
CAP-FXD4700PF+-2%50VPOLYP-FL  
CAP-FXD1000PF+-5%100VCERCOG  
14,15  
14,15  
14,15  
ALL  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
C29  
14,15  
16,17  
ALL  
C29  
C31  
C33,35,44,45,46- 0160-4822  
,56  
ALL  
C36  
0160-7548  
0160-4801  
0160-7673  
0180-4085  
0180-4355  
0180-4437  
0160-4065  
0160-7049  
0160-7363  
0160-4808  
0160-0301  
0811-3478  
0811-3839  
0811-3864  
0811-3861  
1
1
1
1
1
1
1
2
1
1
4
2
2
2
2
6
CAP-FXD2200PF+-10%50VCERY5P  
CAP-FXD100PF+-5%100VCERCOG  
CAP-FXD.047UF+-10%100VPOLYP-MET  
CAP-FXD330UF+-20%35VAL-ELECTLT  
CAP-FXD470UF50V+-20%AL-ELECTLT  
CAP-FXD47UF+-20%50VAL-ELECTLT  
CAP-FXD0.1UF+-20%250VPPR-MET  
CAP-FXD4700PF+-20%250VCERX5V  
CAP-FXD1UF+-10%250VPOLYP-MET  
CAP-FXD470PF+-5%100VCERCOG  
CAP-FXD0.012UF+-10%200VPE-FL  
RESISTOR0.1+-1%5WPWNTC=0+-90  
RESISTOR0.2+-1%5WPWNTC=0+-90  
RESISTOR0.6+-5%5WPWNTC=0+-90  
RESISTOR1.78+-1%5WPWNTC=0+-90  
RESISTOR31.6K+-1%.125WTFTC=0+-100  
ALL  
ALL  
ALL  
ALL  
ALL  
ALL  
ALL  
ALL  
16,17  
16,17  
17  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
C37  
C40  
C52  
C53  
C57  
C58  
C59,60  
C61  
C62  
C63,64,65,66  
R1,2  
R1,2  
R1,2  
R1,2  
14  
15  
16  
17  
R3,7,21,22,105,10 0698-3160  
6
ALL  
R4,71,73,76,112  
R5,6,80  
R8,9,12  
R8,9  
0757-0465  
0757-0401  
0757-0280  
0761-0021  
0699-2715  
0811-1806  
0811-2188  
0764-0007  
0699-3105  
0757-0461  
0698-3157  
0757-0442  
0757-0465  
5
3
3
2
2
1
1
1
1
1
3
1
1
6
RESISTOR100K+-1%.125WTFTC=0+-100  
RESISTOR100+-1%.125WTFTC=0+-100  
RESISTOR1K+-1%.125WTFTC=0+-100  
RESISTOR1K+-5%1WTFTC=0+-100  
RESISTOR-FUSE1OHM+-5%;0.5W@70  
RESISTOR2K+-5%3WPWI20PPM  
ALL  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
ALL  
14,15  
16,17  
ALL  
R10,108  
R11  
14  
R11  
RESISTOR5K+-5%3WPWI20PPM  
15  
R11  
RESISTOR27K+-5%2WMOTC=0+-200PPM  
RESISTOR45K+-5%2WMOTC=0+-500PPM  
RESISTOR68.1K+-1%.125WTFTC=0+-100  
RESISTOR19.6K+-1%.125WTFTC=0+-100  
RESISTOR10K+-1%.125WTFTC=0+-100  
RESISTOR100K+-1%.125WTFTC=0+-100  
RESISTOR1K+-1%.125WTFTC=0+-100  
16  
R11  
17  
R13  
14,15  
14,15  
14,15  
14,15  
ALL  
R14,48,52  
R15  
R16  
R17,18,66,78,110, 0757-0280  
123  
R19,113  
R20,23  
0698-0083  
0757-0463  
2
2
RESISTOR1.96K+-1%.125WTFTC=0+-100  
RESISTOR82.5K+-1%.125WTFTC=0+-100  
ALL  
ALL  
28480  
28480  
A-14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table A-10. Replaceable Parts List (Cont'd)  
Reference Designator  
Agilent Part  
Number  
Q'ty  
Description  
Model  
Mfr. P/N  
Mfr.  
Code  
R24,26,27,37,38, 0757-0442  
64,88,117,120  
9
RESISTOR10K+-1%.125WTFTC=0+-100  
ALL  
28480  
R25,30,33  
R28,111  
0698-8824  
0698-3228  
3
RESISTOR562K+-1%.125WTFTC=0+-100  
RESISTOR49.9K+-1%.125WTFTC=0+-100  
RESISTOR46.4K+-1%.125WTFTC=0+-100  
ALL  
ALL  
ALL  
28480  
28480  
28480  
2
R29,68,86,89,91, 0698-3162  
9-  
11  
2,95,96,99,114,12  
1
R31,32,35  
R34  
0698-0084  
0757-0288  
0698-3518  
0757-0439  
0757-0441  
0698-8580  
0757-0440  
0698-4471  
0698-3498  
0757-0442  
0757-0431  
0698-4438  
0698-0063  
0757-0439  
0698-4473  
0757-0454  
0698-4503  
0757-0467  
0757-0346  
0698-3438  
0757-0293  
0757-0401  
0757-0489  
0698-4123  
0757-0293  
0811-3909  
0811-3909  
0811-4118  
0811-3861  
0811-1799  
0813-0001  
0811-0071  
0811-1808  
0698-0085  
0698-7634  
3
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
4
4
2
4
2
2
2
2
1
1
2
1
1
1
1
1
1
RESISTOR2.15K+-1%.125WTFTC=0+-100  
RESISTOR9.09K+-1%.125WTFTC=0+-100  
RESISTOR7.32K+-1%.125WTFTC=0+-100  
RESISTOR6.81K+-1%.125WTFTC=0+-100  
RESISTOR8.25K+-1%.125WTFTC=0+-100  
RESISTOR9.53K+-1%.125WTFTC=0+-100  
RESISTOR7.5K+-1%.125WTFTC=0+-100  
RESISTOR7.15K+-1%.125WTFTC=0+-100  
RESISTOR8.66K+-1%.125WTFTC=0+-100  
RESISTOR10K+-1%.125WTFTC=0+-100  
RESISTOR2.43K+-1%.125WTFTC=0+-100  
RESISTOR3.09K+-1%.125WTFTC=0+-100  
RESISTOR5.23K+-1%.125WTFTC=0+-100  
RESISTOR6.81K+-1%.125WTFTC=0+-100  
RESISTOR8.06K+-1%.125WTFTC=0+-100  
RESISTOR33.2K+-1%.125WTFTC=0+-100  
RESISTOR66.5K+-1%.125WTFTC=0+-100  
RESISTOR121K+-1%.125WTFTC=0+-100  
RESISTOR10+-1%.125WTFTC=0+-100  
RESISTOR147+-1%.125WTFTC=0+-100  
RESISTOR1.96K+-1%.125WTFTC=0+-100  
RESISTOR100+-1%.125WTFTC=0+-100  
RESISTOR10+-1%.25WTFTC=0+-100  
RESISTOR499+-1%.125WTFTC=0+-100  
RESISTOR1.96K+-1%.125WTFTC=0+-100  
RESISTOR0.2+-1%10WPWNTC=0+-90  
RESISTOR0.2+-1%10WPWNTC=0+-90  
RESISTOR0.6+-1%10WPWNTC=0+-90  
RESISTOR1.78+-1%5WPWNTC=0+-90  
RESISTOR390+-5%3WPWITC=0+-20  
ALL  
ALL  
14  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
R36  
R36  
15  
R36  
16  
R36  
17  
R39  
14  
R39  
15  
R39  
16  
R39  
17  
R40  
14  
R40  
15  
R40  
16  
R40  
17  
R41  
14  
R41  
15  
R41  
16  
R41  
17  
R42,44,54,56  
R42,44,54,56  
R43,55  
R43,45,55,57  
R46,53  
R47,50  
R49,51  
R58,59  
R58  
14,15  
16,17  
16,17  
14,15  
14,15  
14,15  
14,15  
14  
15  
R58  
16  
R58,59  
R60  
17  
14  
R60  
RESISTOR1K+-5%3WPWITC=0+-20  
15  
R60  
RESISTOR1.52K+-5%3WPWITC=0+-20  
RESISTOR2.6K+-5%3WPWITC=0+-20  
RESISTOR2.61K+-1%.125WTFTC=0+-100  
RESISTOR42.2K+-1%.125WTFTC=0+-100  
16  
R60  
17  
R61  
ALL  
14  
R62  
A-15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
R62  
R62  
0698-4514  
0757-0481  
1
1
RESISTOR105K+-1%.125WTFTC=0+-100  
RESISTOR475K+-1%.125WTFTC=0+-100  
Table A-10. Replaceable Parts List (Cont'd)  
15  
16  
28480  
28480  
Reference Designator  
Agilent Part  
Number  
Q'ty  
Description  
Model  
Mfr. P/N  
Mfr.  
Code  
R62  
0698-8826  
0698-8827  
0757-0274  
0757-0438  
2100-4306  
0698-3243  
0698-3459  
0698-3158  
0757-0465  
0757-0289  
0757-0290  
0757-0458  
0698-8123  
0757-0444  
0698-3245  
0698-3136  
0698-3430  
0757-0395  
0698-4767  
0698-3460  
0698-8825  
0698-8827  
0698-3157  
0698-8123  
0757-0461  
0757-0440  
0698-3444  
0757-0346  
0698-3581  
2100-4357  
2100-4305  
0698-3455  
0757-0465  
0757-0461  
0698-3160  
0698-4123  
0698-3441  
0698-3438  
0757-0428  
0698-3156  
1
1
1
1
1
1
1
1
1
1
1
1
1
2
2
2
1
1
1
1
1
1
2
1
1
2
1
1
1
1
1
1
1
1
1
1
1
1
1
2
RESISTOR825K+-1%.125WTFTC=0+-100  
RESISTOR1M+-1%.125WTFTC=0+-100  
RESISTOR1.21K+-1%.125WTFTC=0+-100  
RESISTOR5.11K+-1%.125WTFTC=0+-100  
RESISTOR-TRMR50K10%TKFTOP-ADJ25-T  
RESISTOR178K+-1%.125WTFTC=0+-100  
RESISTOR383K+-1%.125WTFTC=0+-100  
RESISTOR23.7K+-1%.125WTFTC=0+-100  
RESISTOR100K+-1%.125WTFTC=0+-100  
RESISTOR13.3K+-1%.125WTFTC=0+-100  
RESISTOR6.19K+-1%.125WTFTC=0+-100  
RESISTOR51.1K+-1%.125WTFTC=0+-100  
RESISTOR26.1K+-1%.125WTFTC=0+-100  
RESISTOR12.1K+-1%.125WTFTC=0+-100  
RESISTOR20.5K+-1%.125WTFTC=0+-100  
RESISTOR17.8K+-1%.125WTFTC=0+-100  
RESISTOR21.5+-1%.125WTFTC=0+-100  
RESISTOR56.2+-1%.125WTFTC=0+-100  
RESISTOR147K+-1%.125WTFTC=0+-100  
RESISTOR422K+-1%.125WTFTC=0+-100  
RESISTOR681K+-1%.125WTFTC=0+-100  
RESISTOR1M+-1%.125WTFTC=0+-100  
RESISTOR19.6K+-1%.125WTFTC=0+-100  
RESISTOR26.1K+-1%.125WTFTC=0+-100  
RESISTOR68.1K+-1%.125WTFTC=0+-100  
RESISTOR7.5K+-1%.125WTFTC=0+-100  
RESISTOR316+-1%.125WTFTC=0+-100  
RESISTOR10+-1%.125WTFTC=0+-100  
RESISTOR13.7K+-1%.125WTFTC=0+-100  
RESISTOR-VAR10K+-10%  
17  
28480  
28480  
28480  
28480  
32997  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
R63  
ALL  
ALL  
ALL  
ALL  
14  
R65  
R67  
R69  
3296Y-1-503  
R70  
R70  
15  
R70  
16  
R70  
17  
R72  
14  
R72  
15  
R72  
16  
R72  
17  
R74,75  
R74,75  
R74,75  
R77  
14,15  
16  
17  
14,15  
16,17  
14  
R77  
R79  
R79  
15  
R79  
16  
R79  
17  
R81,122  
R82  
ALL  
14,15  
16,17  
ALL  
ALL  
ALL  
ALL  
ALL  
ALL  
14  
R82  
R83,119  
R87  
R90  
R93  
R94  
52UAL-T22-A15 32997  
R97  
RESISTOR-TRMR10K10%TKFTOP-ADJ25-T  
RESISTOR261K+-1%.125WTFTC=0+-100  
RESISTOR100K+-1%.125WTFTC=0+-100  
RESISTOR68.1K+-1%.125WTFTC=0+-100  
RESISTOR31.6K+-1%.125WTFTC=0+-100  
RESISTOR499+-1%.125WTFTC=0+-100  
RESISTOR215+-1%.125WTFTC=0+-100  
RESISTOR147+-1%.125WTFTC=0+-100  
RESISTOR1.62K+-1%.125WTFTC=0+-100  
RESISTOR14.7K+-1%.125WTFTC=+-100  
3296Y-1-103  
32997  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
R98  
R98  
15  
R98  
16  
R98  
17  
R100  
R101  
R102  
R103  
R104,107  
ALL  
ALL  
ALL  
ALL  
ALL  
A-16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
R109  
R115  
0698-3153  
0757-0462  
1
1
RESISTOR3.83K+-1%.125WTFTC=0+-100  
RESISTOR75K+-1%.125WTFTC=0+-100  
Table A-10. Replaceable Parts List (Cont'd)  
ALL  
14  
28480  
28480  
Reference Designator  
Agilent Part  
Number  
Q'ty  
Description  
Model  
Mfr. P/N  
Mfr.  
Code  
R115  
0757-0463  
0757-0462  
0757-0461  
0698-4489  
0698-3161  
0698-4494  
0698-8678  
0698-8812  
0698-8825  
0757-0401  
1826-0144  
1826-0346  
1826-0412  
1826-0138  
1826-0665  
1826-1297  
1990-1659  
1826-1702  
1826-1075  
1826-0468  
1826-0393  
1826-0221  
1826-0346  
1820-1197  
1990-1659  
1855-0989  
1855-0536  
1854-0477  
1854-0477  
1853-0281  
1853-0041  
1901-1273  
1906-0284  
1901-0033  
1
1
1
1
1
1
1
1
1
1
1
1
3
2
2
3
1
1
1
1
1
1
1
1
4
2
2
4
4
2
1
2
1
5
6
5
2
1
RESISTOR82.5K+-1%.125WTFTC=0+-100  
RESISTOR75K+-1%.125WTFTC=0+-100  
RESISTOR68.1K+-1%.125WTFTC=0+-100  
RESISTOR28K+-1%.125WTFTC=0+-100  
RESISTOR38.3K+-1%.125WTFTC=0+-100  
RESISTOR35.7K+-1%.125WTFTC=0+-100  
RESISTOR178+-1%.125WTFTC=0+-100  
RESISTOR1+-1%.125WTFTC=0+-100  
RESISTOR681K+-1%.125WTFTC=0+-100  
RESISTOR100+-1%.125WTFTC=0+-100  
ICVRGLTR-FXD-POS4.8/5.2VTO-220PKG  
ICOPAMPGPDUAL8PINDIP-P  
15  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
04713  
27014  
27014  
27014  
27014  
27014  
14936  
28480  
27014  
04713  
27014  
04713  
27014  
01295  
14936  
28480  
28480  
04713  
04713  
04713  
04713  
14936  
14936  
27014  
14936  
04713  
14936  
14936  
R115  
16  
R115  
17  
R116  
14  
R116  
15,16  
17  
R116  
R118  
ALL  
16,17  
16,17  
ALL  
ALL  
R124  
R125  
R126  
U1  
MC7805CT  
U2  
14,15 LM358N  
U3,16,20  
U4,5  
ICCOMPARATORPRCNDUAL8PINDIP-P  
ICCOMPARATORGPQUAD14PINDIP-P  
ICOPAMPLOW-BIAS-H-IMPDQUAD14PIN  
ALL  
ALL  
ALL  
LM393N  
LM339  
U6,9  
LF347BN  
LM336BZ-5.0  
MCP3020Z  
U7,15,17  
U8  
ICVRGLTR-V-REF-FXD4.8/5.2VTO-92PKG ALL  
OPTO-ISOLATORLED-TRIACIF=100MA-MAX  
ICOPAMPPRCN8PINDIP-P  
ALL  
ALL  
ALL  
ALL  
ALL  
U10  
U11  
ICOPAMPGPDUAL8PINDIP-P  
LF442CN  
MC3423P1  
LM317T  
U12  
ICVRGLTR-OV-V-SEN2.3/37.8V8-DIP-P  
ICVRGLTR-ADJ-POS1.2/37VTO-220PKG  
U13  
U14  
ICVRGLTR-FXD-NEG-11.5/-12.5VTO-220 ALL  
MC7912CT  
LM358N  
U18  
ICOPAMPGPDUAL8PINDIP-P  
ALL  
ALL  
U19  
ICGATETTL/LSNANDQUAD2-INP  
OPTO-ISOLATORLED-TRIACIF=100MA-MAX  
SN74LS00N  
U21,22,23,24  
Q1,4  
16,17 MCP3020Z  
TRANSISTORMOSFETN-CHANE-MODETO247AC 14,15 IRFP044  
TRANSISTORMOSFETN-CHANE-MODETO-3SI 16,17  
Q1,4  
Q2,3,10,11  
Q5,6,7,8  
Q6,7  
TRANSISTORNPN2N2222ASITO-18PD=500MW ALL  
2N2222A  
TRANSISTORNPN2N2222ASITO-18PD=500MW 14,15 2N2222A  
TRANSISTORPNP2N2907ASITO-18PD=400MW 16,17 2N2907A  
Q9  
TRANSISTORPNPSITO-39PD=1WFT=60MHZ  
DIODE-PWRRECT100V6A35NS  
DIODE-FWBRIDGE100V1A  
ALL  
ALL  
ALL  
MM5007  
FE6B  
CR1,19  
CR2  
DF01  
CR3,4,5,6,7  
DIODE-GENPRP180V200MADO-35  
DIODE-PWRRECT400V1A50NSDO-41  
THRYSTER-SCRTO-220ABVRRM=200V  
DIODE-PWRRECT400V1A50NSDO-41  
DIODE-BRIDGE600V6A  
14,15 1N645  
CR8,9,16,17,31,32 1901-1149  
CR10,12,15,18,20 1884-0332  
ALL  
ALL  
UF4004  
MCR264-4  
CR11,14  
CR13  
1901-1149  
1906-0400  
14,15 UF4004  
ALL GBU8J  
A-17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table A-10. Replaceable Parts List (Cont'd)  
Reference Designator  
Agilent Part  
Number  
Q'ty  
Description  
Model  
Mfr. P/N  
Mfr.  
Code  
CR21,22,23,24,25, 1901-0033  
26,27,28,29,30  
10  
DIODE-GENPRP180V200MADO-35  
ALL  
1N645  
27014  
VR1,2,3  
RT1,2  
C67  
1902-0579  
0837-0261  
0160-0263  
3
2
1
DIODE-ZNR5.1V5%PD=1WIR=10UA  
DIODE-VARISTOR  
ALL  
ALL  
ALL  
1N4733APL  
V275LA20A  
04713  
34371  
CAP-FXD0.22uF+-20%50VCER  
MAGNETIC DEVICE  
T1  
9100-5068  
9100-5069  
NOP/N  
1
1
1
1
2
1
TRANSFORMER-POWER  
TRANSFORMER-POWER  
TRANSFORMER-POWER  
TRANSFORMER-POWER  
TRANSFORMER-PULSE;PRIIND:5MH  
CORE-SHIELDINGBEAD  
14  
28480  
28480  
28480  
28480  
28480  
28480  
T1  
15  
T1  
16  
T1  
9100-5070  
9100-4969  
9170-0894  
17  
T2,3  
L1  
14,15  
ALL  
MISCELLANEOUS  
J1  
1252-4159  
1
CONNECTOR-POSTTYPE2.5-PIN-SPCG11-CO- ALL  
NT  
28480  
TP1-16  
0360-2359  
3101-3237  
3101-3238  
3101-1914  
3101-3115  
3101-3083  
0360-2548  
0360-2545  
0360-2546  
0360-2547  
16  
4
TERMINAL-TESTPOINT.230INABOVE  
SWITCH-SLSPDTSUBMIN6A250VAC  
SWITCH-SLDPDTSUBMIN6A250VAC  
SWITCH-SL2-DPDTSTD1.5A250VACPC  
SWITCH-PBMOM.5A250VAC  
SWITCH-PBDPSTALTNG6A250VAC  
TERMINALBLOCK  
ALL  
ALL  
ALL  
ALL  
ALL  
ALL  
ALL  
ALL  
ALL  
ALL  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
28480  
S1,2,3,4  
S5  
S6  
S7  
S8  
1
1
1
1
3
1
TERMINALBLOCK  
9
TERMINALBLOCK  
1
TERMINALBLOCK  
A-18  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table A-11. Component Value  
Component  
Model  
E3614A  
E3615A  
E3616A  
E3617A  
C2  
C7,8  
1000UF25V+-20%AL-ELECTLT  
39000UF25V+-20%AL-ELECTLT  
470UF50V+-20%AL-ELECTLT  
12000UF63V+-20%AL-ELECTLT  
0.01UF100V+-10%CERX7R  
0.1UF50V+-10%CERX7R  
0.1UF630V+-10%POLYE-FL  
1000PF100V+-5%COG  
330UF50V+-20%AL-ELECTLT  
5600UF100V+-20%AL-ELECTLT  
220UF100V+-20%AL-ELECTLT  
2700UF160V+-20%AL-ELECTLT  
C11,16,17,25,26 0.01UF100V+-10%CERX7R  
C12  
0.1UF50V+-10%CERX7R  
0.1UF630V+-10%POLYE-FL  
1000PF100V+-5%COG  
C18,21,24,27  
C19,22  
C20,23  
C29  
1UF50V+-20%AL-ELECTLT  
0.01UF100V+-10%CERX7R  
1UF50V+-20%AL-ELECTLT  
0.01UF100V+-10%CERX7R  
0.047UF100V+-10%POLYP-MET  
1UF+-10%250VPOLYP-MET  
470PF+-5% 100VCERCOG  
0.047UF100V+-10%POLYP-MET  
1UF+-10%250VPOLYP-MET  
470PF+-5% 100VCERCOG  
0.012UF+-10%200VPE-FL  
1.78+-1%5W  
C61  
C62  
C63,64,65,66  
R1,2  
0.1+-1%5W  
0.2+-1%5W  
0.6+-5%5W  
1K+-5%1W  
27k+-5%2W  
R8,9  
1K1%.125W  
1K1%.125W  
1K+-5%1W  
R11  
2K+-5%3W  
5K+-5%3W  
45K+-5%2W  
R12  
1K+-1%.125W  
68.1K+-1%.125W  
19.6K+-1%.125W  
10K+-1%.125W  
100K+-1%.125W  
7.32K+-1%.125W  
7.5K+-1%.125W  
2.43K+-1%.125W  
8.06K+-1%.125W  
10+-1%.125W  
100+-1%.125W  
100+-1%.125W  
10+-1%.25W  
1K+-1%.125W  
68.1K+-1%.125W  
19.6K+-1%.125W  
10K+-1%.125W  
100K+-1%.125W  
6.81K+-1%.125W  
7.15K+-1%.125W  
3.09K+-1%.125W  
33.2K+-1%.125W  
10+-1%.125W  
100+-1%.125W  
100+-1%.125W  
10+-1%.25W  
R13  
R14,48,52  
R15  
R16  
R36  
8.25K+-1%.125W  
8.66K+-1%.125W  
5.23K+-1%.125W  
66.5K+-1%.125W  
147+-1%.125W  
1K+-1%.125W  
9.53K+-1%.125W  
10K+-1%.125W  
6.81K+-1%.125W  
121K+-1%.125W  
147+-1%.125W  
1K+-1%.125W  
R39  
R40  
R41  
R42,44,54,56  
R43,55  
R45,57  
R46,53  
R47,50  
R49,51  
R58  
499+-1%.125W  
1.96K+-1%.125W  
0.2+-1%10W  
499+-1%.125W  
1.96K+-1%.125W  
0.2+-1%10W  
0.6+-1%10W  
1.78+-1%5W  
R59  
0.2+-1%10W  
1.78+-1%5W  
R60  
390+-5%3W  
1K+-5%3W  
1.52K+-5%3W  
2.6K+-5%3W  
R62  
42.2K+-1%.125W  
178K+-1%.125W  
13.3K+-1%.125W  
12.1K+-1%.125W  
21.5+-1%.125W  
147K+-1%.125W  
26.1K+-1%.125W  
261K+-1%.125W  
75K+-1%.125W  
28K+-1%.125W  
105K+-1%.125W  
383K+-1%.125W  
6.19K+-1%.125W  
12.1K+-1%.125W  
21.5+-1%.125W  
422K+-1%.125W  
26.1K+-1%.125W  
100K+-1%.125W  
82.5K+-1%.125W  
38.3K+-1%.125W  
475K+-1%.125W  
23.7K+-1%.125W  
51.1K+-1%.125W  
20.5+-1%.125W  
56.2+-1%.125W  
681K+-1%.125W  
68.1K+-1%.125W  
68.1K+-1%.125W  
75K+-1%.125W  
825K+-1%.125W  
100K+-1%.125W  
26.1K+-1%.125W  
17.8K+-1%.125W  
56.2+-1%.125W  
1M+-1%.125W  
R70  
R72  
R74,75  
R77  
R79  
R82  
68.1K+-1%.125W  
31.6K+-1%.125W  
68.1K+-1%.125W  
35.7K+-1%.125W  
1+-1%.125W  
R98  
R115  
R116  
38.3K+-1%.125W  
1+-1%.125W  
R124  
R125  
681K+-1%.125W  
MOSFETN-CHANE-MODETO3SI  
681K+-1%.125W  
MOSFETN-CHANE-MODETO3SI  
Q1,4  
MOSFETN-CHANE-MODETO204AE  
NPN2N2222ASITO-18PD=500MW2  
NPN2N2222ASITO-18PD=500MW2  
ICOPAMPGPDUAL8PINDIP-P  
MOSFETN-CHANE-MODETO204AE  
NPN2N2222ASITO-18PD=500MW  
NPN2N2222ASITO-18PD=500MW2  
ICOPAMPGPDUAL8PINDIP-P  
Q5,8  
Q6,7  
PNP2N2907ASITO-18PD=400MW  
OPTO-ISOLATORLED-TRIAC  
PNP2N2907ASITO-18PD=400MW  
OPTO-ISOLATORLED-TRIAC  
U2  
U21,22,23,24  
CR3,4,5,6,7  
CR11,14  
T1  
DIODE-GENPRP180V200MAD0-35  
DIODE-GENPRP180V200MAD0-35  
DIODE-PWRRECT400V1A50NSDO-41 DIODE-PWRRECT400V1A50NSDO-41  
TRANSFORMER-POWERFORE3614A TRANSFORMER-POWERFORE3615A  
TRANSFORMER-POWERFORE3616A  
TRANSFORMER-POWERFORE3617A  
T2,3  
TRANSFORMER-PULSE;PRIIND:5MH TRANSFORMER-PULSE;PRIIND:5MH  
A-19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Manual Supplement  
Supplement Agilent Part Number : 5959-5336, Edition 4  
Supplement Print Date : 14 April, 2000  
This supplement updates the following document:  
Agilent E361XA 60W Series Lab Bench DC Power Supplies  
Manual Agilent Part Number : 5959-5310  
What is a manual supplement?  
A manual supplement keeps your manual up-to-date. The supplement, which  
consists of additional pages for your manual, is shipped with the manual that it  
updates. Additional pages have page numbers with a lower-case letter. For  
example, if one additional page is added between pages 1-10 and 1-11, it will be  
numbered 1-10-1.  
This supplement is new information that was not described in the manual  
for remote programming of the E3614A/E3615A/E3616A/E3617A with a voltage  
or current source and resistors.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Voltage and Current Programming of the E3614A/15A/16A/  
17A with a Voltage and Current Source  
Remote analog voltage programming permits control of the regulated output voltage  
or current by means of a remotely varied voltage or current. The stability of the  
programming voltages directly affects the stability of the output. The voltage control  
or current control on the front panel are disabled during analog programming.  
N O T E  
The CV(-) terminal on the rear panel is internally connected to the plus output  
terminal. In following connections, it is recommended to use Figure 2, Figure 4, or  
Figure 6 if the negative terminal of the “Programming Voltage” is not floted from  
its circuits.  
Constant Voltage Mode  
The programming voltage is not isolated from the power supply output. The power  
supply may be programmed with a voltage that is common to either the plus output,  
or the minus output.  
Programming Voltage Common to the Plus output  
Figure 1  
Set the CV switch down on the rear panel, and all others up.  
Vin = 1/A x Vout  
Vout = A x Vin  
Where  
Vout is the power supply output voltage.  
Vin is the programming voltage.  
A is the gain factor and the values of each model are as below.  
Model  
E3614A  
A
1/A  
0.8  
2.0  
3.5  
6.0  
1.25  
0.5  
E3515A  
E3616A  
E3617A  
0.29  
0.17  
1-10-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Programming Voltage Common to the Minus Output  
Figure 2  
Set the CV switch down on the rear panel, and all others up.  
Vin = 1/A x Vout  
Vout = A x Vin  
Where  
Vout is the power supply output voltage.  
Vin is the programming voltage.  
A is the gain factor and the values of each model are as below.  
Model  
E3614A  
A
1/A  
0.44  
0.67  
0.78  
0.86  
2.25  
1.5  
E3515A  
E3616A  
E3617A  
1.29  
1.17  
Alternative Voltage Programming Using Resistors  
Programming Voltage Common to the Plus Output  
Figure 3  
The M/S2 switch must be in the down position. For best results, place a 0.1µF capacitor in  
parallel with R2.  
Vin = (R1/R2) x Vout  
Vout = (R2/R1) x Vin  
Where  
Vout is the power supply output voltage.  
Vin is the programming voltage.  
R1 and R2 should be in the 1Kto 100Krange.  
1-10-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Programming Voltage Common to the Minus Output  
Figure 4  
The output will always be the same or less than the programming voltage.  
The M/S2 switch must be in the down position. For best results, place a 0.1µF capacitor  
in parallel with R2.  
Vin = (R1R2) / R2 x Vout  
Vout = R2 / R1+R2) x Vin  
Where  
Vout is the power supply output voltage.  
Vin is the programming voltage.  
R1 and R2 should be in the 1Kto 100Krange.  
1-10-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Constant Current Mode  
The E3614A/15A/16A/17A may be programmed for constant current with an analog  
voltage or current. Constant current with analog voltage programming can only be  
achieved with a voltage source that is common with the positive output terminal.  
Constant Current with Voltage Programming  
Figure 5  
Set the CC switch down the rear panel, and all others up.  
Vin = 1/A x Iout  
Iout = A x Vin  
Where  
Iout is the power supply output current.  
Vin is the programming voltage.  
A is the transconductance in Amp/Volt and the values of each  
model are as below.  
Model  
E3614A  
A (A/V)  
1/A (V/A)  
1.67  
0.6  
0.3  
E3515A  
E3616A  
E3617A  
3.33  
6.0  
10  
0.17  
0.1  
Constant Current with Current Programming  
When using current to program the power supply, the source must have a  
dynamic range of 10 volts when the programming source is common to the plus  
output and 10 volts plus the maximum output voltage expected when the  
programming source is common to the minus output of the power supply.  
The load to the power supply must be stable for the constant current  
output to be accurate. Current transient response is not specified,  
and depends on the change of the output voltage of the power supply.  
1-10-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Figure 6  
Set the CC switch down, and all others up.  
Iin = 1/A x Iout  
Iout = A x Vin  
Where  
Iout is the power supply output current in amps.  
Iin is the programming current in µamps.  
A is the gain.  
Model  
E3614A  
A (A/µA)  
0.055  
1/A (µV/A)  
18  
E3515A  
E3616A  
E3617A  
0.0278  
0.0158  
0.00928  
35.9  
63.4  
108  
Programming currents can be increased by adding a resistor across the CC+ and CC-. A  
10 volts drop across R1 represents full scale current of the power supply. When a 1 kohm  
resistor is added across R1, the programming currents are as follows with the  
programming current in mA.  
Parallel resistor required for a  
Model  
E3614A  
A (A/mV)  
0.594  
1/A (mA/A)  
1.69  
1 amp/mamp value of A (kohm)  
1.7  
E3515A  
E3616A  
E3617A  
0.297  
0.168  
0.0989  
3.37  
5.95  
1.01  
3.45  
6.28  
11.2  
Current Monitoring  
Current of the power supply can be monitored across the internal current monitoring  
resistor. One side of the resistor is at the +output and A3; the other side of the resistor  
is at A1. The table below shows the resistor value and conversion factors. To obtain  
the current divide the measured voltage by the resistor value or multiply the amps/V  
times the voltage measured.  
Resistor  
Model  
E3614A  
amps/V  
value ()  
0.1  
10  
E3515A  
E3616A  
E3617A  
0.2  
5
0.6  
1.67  
1.12  
0.89  
1-10-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Voltage and Current Programming of the  
E3614A/15A/16A/17A with Resistors  
Remote programming with resistors permits control of the regulated output or current  
by means of a remotely varied resistor. The sum of the resistance of external  
programming resistors (R1 + R2) should be more than 40 kohm. To have more precise  
output voltage, use a variable resistor more than 40 kohm. The voltage control on the  
front panel is disabled during remote resistor programming.  
N O T E  
Do not operate the power supply simultaneously in the remote analog voltage  
programming and in the remote resistor programming.  
Remote Resistor Programming Connections  
Remote resistor programming requires changing the setting of the switches and  
connecting external resistors between “+” and “`-” terminals of “CV” and “VREF”  
terminal or “+” and “-” terminals of “CC” and “VREF” terminal. Any noise picked up  
on the programming leads will appear on the power supply's output and may degrade  
regulation. To reduce noise pickup, use a twisted or shielded pair of wires for  
programming, with the shield grounded at one end only.  
Remote Resistor Programming, Constant Voltage  
Figure 7  
Set the CV switch down on the rear panel, and all others up.  
V
out = A x [VREF x {R/(R + R2 + 100)}]  
Where out is the power supply output voltage.  
A is the gain factor and the values of each model are as below.  
REF is between 10.11 V and 11.40 V.  
V
V
R = (92800 x R1)/(92800 + R1)  
R1 + R2 > 40 kohm  
Model  
E3614A  
A
0.8  
2.0  
3.5  
6.0  
E3515A  
E3616A  
E3617A  
1-10-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Remote Resistor Programming, Constant Current  
Figure 8  
Set the CC switch down on the rear panel, and all others up.  
I
out = A x [VREF x {R/(R + R2 + 100)}]  
Where out is the power supply output current.  
A is the gain factor and the values of each model are as below.  
REF is between 10.11 V and 11.40 V.  
I
V
R = (92800 x R1)/(92800 + R1)  
R1 + R2>> 40 kohm  
Model  
E3614A  
A
0.6  
E3515A  
E3616A  
E3617A  
0.3  
0.17  
0.1  
1-10-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Iꢀ  
CERTIFICATION  
Agilent Technologies certifies that this product met its published specifications at time of shipment from the factory. Agilent  
Technologies further certifies that its calibration measurements are traceable to the United States National Institute of Stan-  
dards and Technology (formerly National Bureau of Standards), to the extent allowed by that organization's calibration facility,  
and to the calibration facilities of other International Standards Organization members.  
WARRANTY  
This Agilent Technologies hardware product is warranted against defects in material and workmanship for a period of three  
years from date of delivery. Agilent software and firmware products, which are designated by Agilent for use with a hardware  
product and when properly installed on that hardware product, are warranted not to fail to execute their programming instruc-  
tions due to defects in material and workmanship for a period of 90 days from date of delivery. During the warranty period,  
either Agilent or Agilent Technologies will, at its option, either repair or replace products which prove to be defective. Agilent  
does not warrant that operation the software, firmware, or hardware shall be uninterrupted or error free.  
For warranty service, with the exception of warranty options, this product must be returned to a service facility designated by  
Agilent. Return to Englewood Colorado Service Center for repair in United States(1-800-258-5165). Customer shall prepay  
shipping charges by (and shall pay all duty and taxes) for products returned to Agilent for warranty service. Except for the  
products returned to Customer from another country, Agilent shall pay for return of products to Customer.  
Warranty services outside the country of initial purchase are included in Agilent's product price, only if Customer pays Agilent  
international prices (defined as destination local currency price, or U.S. or Geneva Export price).  
If Agilent is unable, within a reasonable time, to repair or replace any product to condition as warranted, the Customer shall  
be entitled to a refund of the purchase price upon return of the product to Agilent.  
The warranty period begins on the date of delivery or on the date of installation if installed by Agilent.  
LIMITATION OF WARRANTY  
The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by the Customer, Cus-  
tomer-supplied software or interfacing, unauthorized modification or misuse, operation outside of the environmental specifica-  
tions for the product, or improper site preparation and maintenance. TO THE EXTENT ALLOWED BY LOCAL LAW, NO  
OTHER WARRANTY IS EXPRESSED OR IMPLIED. AND AGILENT SPECIFICALLY DISCLAIMS THE IMPLIED WARRAN-  
TIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.  
For consumer transactions in Australia and New Zealand:  
The warranty terms contained in this statement, except to the extent lawfully permitted, do not exclude, restrict or modify and  
are in addition to the mandatory rights applicable to the sale of this product to you.  
EXCLUSIVE REMEDIES  
TO THE EXTENT ALLOWED BY LOCAL LAW, THE REMEDIES PROVIDED HEREIN ARE THE CUSTOMER'S SOLE AND  
EXCLUSIVE REMEDIES. AGILENT SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR  
CONSEQUENTIAL DAMAGES, WHETHER BASED ON CONTRACT, TORT, OR ANY OTHER LEGAL THEORY.  
ASSISTANCE  
The above statements apply only to the standard product warranty. Warranty options, extended support contacts, product  
maintenance agreements and customer assistance agreements are also available. Contact your nearest Agilent Technolo-  
gies Sales and Service office for further information on Agilent's full line of Support Programs.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
DECLARATION OF CONFORMITY  
according to ISO/IEC Guide 22 and EN 45014  
Manufacturer’s Name:  
Agilent Technologies, Inc.  
Manufacturer’s Address:  
345-15, Kasan-dong, Kumchon-ku,  
Seoul 153-023 Korea  
declares, that the products  
Product Name:  
DC Power Supply  
Model Numbers:  
E3614A, E3615A, E3614A, E3615A, and E3617A  
All Options  
Product Options:  
conforms to the following Product Specifications:  
Safety:  
EMC:  
IEC 1010-1:1990+A1:1992 / EN 61010-1:1993  
1)  
CISPR 11:1990 / EN 55011:1991 - Group 1 Class A  
EN 50082-1:1992  
IEC 801-2 : 1991 - 4KV CD, 8KV AD  
IEC 801-3 : 1984 - 3V/m  
IEC 801-4 : 1988 - 1kV Power Lines  
0.5kV Signal Lines  
Supplementary Information: The product herewith comply with the requirements of the  
Low Voltage Directive 73/23/EEC and the EMC Directive 89/336/EEC and carry the “CE”  
mark accordingly.  
1)  
The products was tested in a typical configuration with Agilent Technologies Test System.  
Seoul, Korea  
November 1, 1999  
Quality Manager  
European Contact for regulatory topics only: Hewlett-Packard GmbH, HQ-TRE, Herrenberger Strabe 110-140,  
D-71034 Böbligen (FAX: +49-7031-143143).  
Download from Www.Somanuals.com. All Manuals Search And Download.  

2nd Ave Coffeemaker X User Manual
3Com Switch DUA1770 0AAA04 User Manual
Ace Laptop LP Series User Manual
Aiphone Car Video System MFH U User Manual
Aiphone Intercom System IS SS HID User Manual
Aiwa CD Player XP ER800N User Manual
American Standard Outdoor Shower T675500 User Manual
AO Smith Water Heater KGA SG 30 User Manual
Asus Computer Accessories USBBT211 User Manual
Beko Refrigerator D1 8450 SM User Manual