Agilent Technologies Network Hardware 85107B User Manual

Agilent  
Upgrade Guide for the 8510  
Vector Network Analyzer  
Product Note  
85107B, 45 MHz to 50 GHz in coax  
85106D with option 001,  
45 MHz to 50 GHz in coax,  
above 50 GHz in waveguide  
8510XF  
on-wafer configuration  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Vector network analyzer (VNA) overview  
Vector network analyzer (VNA) measurement systems are used to fully  
characterize the linear behavior of two port devices or networks. Device  
characteristics include the magnitude and phase data of the transmission or  
reflection parameters that are required to determine complex impedance, both  
resistive and reactive components, shown in Figure 1. A network’s behavior is  
linear when (1) a linear change in the input results in a linear change in the  
output, and (2) the output, resulting from multiple input signals, is the same  
as the sum of the outputs resulting from independent input signals. Some  
examples of linear networks are filters, amplifiers, cables and isolators.  
A network analyzer measurement system can be divided into four major parts  
shown in Figure 2:  
1. A signal source providing the incident signal  
2. Signal separation devices to separate the incident, reflected and trans-  
mitted signals, and then down converts the microwave signals to a lower  
intermediate (IF) signal  
3. A receiver to attain the IF signal and down convert it to DC  
4. A signal processor and display section that processes the data and displays  
information on a CRT  
Incident  
DUT  
Transmitted  
Reflected  
Reflection  
(Reflected/incident):  
Input SWR  
Transmission  
(Transmitted/incident):  
Gain/loss  
Return loss  
Isolation  
Input/output impedance  
Reflection coefficient  
S-Parameters S11 , S22  
Insertion phase  
Group delay  
S-Parameters S21  
, S  
12  
Figure 1. Device characterization  
Incident  
DUT  
Transmitted  
Reflected  
Source  
Signal separation  
Reflected  
Transmitted  
Incident  
Receiver/Detector  
Processor/Display  
Figure 2. Network analyzer configuration  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
The Agilent Technologies 8510 network analyzer consists of a family of  
compatible products where each part is a separate system component. Each  
complete system includes the 8510 network analyzer, a test set, and compatible  
sources (measurement accessories are also needed to complete the measure-  
ment setup), shown in Figure 3. The frequency range of the measurement sys-  
tem is determined by the test set or the compatible source’s lesser frequency.  
Example: if a 50 GHz test set is used with a 20 GHz source, the maximum  
frequency measured is 20 GHz.  
8510  
Receiver/detector  
network  
processor/display  
analyzer  
+
8510  
Signal  
separation  
Test set  
851X  
VNA  
=
system  
+
Source  
8360  
Stimulus  
Figure 3. 8510 measurement systems are made up of modular components, the 8510  
network analyzer, a test set and a compatible source.  
4
Download from Www.Somanuals.com. All Manuals Search And Download.  
Agilent 8510 family of products  
85106D  
+
U-band  
V-band  
Waveguide  
Waveguide  
systems  
Test set modules  
Q-band  
W-band  
Combined coax &  
8510XF  
85107B  
Single-sweep  
high  
performance  
systems  
8510SX  
8510E  
85108A (50 GHZ)  
85108A*  
85108L  
50 GHz  
75 GHz  
110 GHz  
20 GHz  
45 MHz  
2 GHz 26.5 GHz  
* 85108A also available in frequency range of 45 MHz to 20 GHz  
Figure 4. 8510 family of system solutions  
Agilent 8510 network analyzer  
Tables 1 and 2 show the different features of the 8510A, B, and C, hardware  
and firmware, respectively.  
Table 1. Hardware revision and measurement capability  
Product  
8510A  
8510B  
8510C  
Firmware revisions  
A.02.00  
Hardware features  
Tape drive  
B.03.00 to B.06.54  
C.06.00 to present  
Tape drive  
• Large color display  
• Internal disk drive  
(LIF or DOS file formats)  
• 4 S-parameter display  
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table 2. Firmware revision and measurement capability  
Revision  
Time domain1  
Pulse2  
Power domain Other added key features  
limit lines  
A.02.00  
B.03.00  
• Adapter removal calibration  
• Direct control of two sources  
B.03.11  
B.04.00  
• Direct control of up to 4 test sets  
• 40 GHz test set capability  
• Frequency subset calibration  
B.05.00  
B.05.11  
B.06.30  
B.06.54  
C.06.00  
• Lightwave data format  
• 50 GHz source/test set compatibility  
• Color display  
• Support for hardcopy printer plots  
C.06.54  
C.07.00  
C.07.01  
C.07.14  
C.07.16  
C.08.10  
• Color printer capability  
• Connector compensation  
• Y2K compliance  
• CRT  
• LCD  
Agilent 8510 Firmware upgrade kits  
Figure 5 diagrams which product to order (numbers shown above arrows)  
to upgrade the existing firmware. For example, to upgrade Rev. C.06.54 to  
Rev. C.07.XX, order 11575J.  
8510C  
Rev. C.06.00  
Rev. C.06.50  
Rev. C.06.54  
11575J  
11575J  
Rev. C.07.XX  
Note: Upgrade kit 11575J has Option 010 (time domain) available.  
If you currently have time domain, you must order Option 010.  
Figure 5. 8510 firmware upgrade kits  
1. Time domain capabilities require Option 010 on the 8510. Time domain option can be upgraded by ordering  
85012C for 8510C.  
2. Pulse capabilities require Option 008 on the 8510 working with either 85110L or 85110A pulsed-RF S-parameter  
test set. Pulse option on the 8510 can be upgraded by ordering 85111B for 8510C. Pulse capability is not available  
with 8510A or 8510B.  
6
Download from Www.Somanuals.com. All Manuals Search And Download.  
Test Sets  
Table 3. Family of test sets  
The test set in a VNA measurement  
system is used to separate the inci-  
dent, the reflected, and the transmitted  
signals. It is also used to convert  
the RF signal to IF (intermediate  
fre- quency) signal and to pass the IF  
to the receiver.  
Products  
Frequency  
range (GHz) (application)  
type needed  
Test set  
Recommend  
RF source1  
needed  
Recommend  
LO source1  
(GHz)  
Test port  
connector2  
(GHz)  
8511A3  
8511B3  
8512A  
8513A  
0.045 to 26.5 Frequency  
Converter  
N/A  
N/A  
3.5 mm (M)  
0.045 to 50  
0.400 to 18  
Frequency  
Converter  
N/A  
N/A  
2.4 mm (F)  
Once signals are separated, their  
individual magnitude and phase  
differences can be measured. Test  
sets are classified into two groups: (1)  
one-path transmission or reflection  
and (2) two-path S-parameter allow-  
ing forward and reverse measure-  
ments of a two port device with a  
single connection. S-parameter test  
sets can be divided into two types: (1)  
sampler- based and (2) mixer-based,  
shown in Table 3. Sampler-based test  
sets require one external source to  
provide the RF stimulus. The mixer-  
based test sets require two external  
sources; one to provide the RF stimu-  
lus and the other to provide the LO  
signal.  
Transmission/  
Reflection  
Obsolete  
Obsolete  
0.045 to 26.5 Transmission/  
Reflection  
8514A  
8514B  
0.500 to 18  
0.045 to 20  
S-parameter  
Obsolete  
-----  
S-parameter  
83621B  
83631B  
3.5 mm (M)  
(0.045 to 20)  
Sampler-based  
8515A  
0.045 to 26.5 S-parameter  
-----  
-----  
3.5 mm (M)  
Sampler-based (0.045 to 26.5)  
8516A  
8517A  
8517B  
0.045 to 40  
0.045 to 50  
0.045 to 50  
S-parameter  
S-parameter  
Obsolete --- replaced with 8517A  
Obsolete --- replaced with 8517B  
S-parameter  
83651B  
-----  
2.4 mm (M)  
Sampler-based (0.045 to 50)  
85110L  
85110A  
0.045 to 2  
2 to 20  
Pulsed-RF  
S-parameter  
Mixer-based  
83620B  
#001, 004,  
008, H80  
83620B  
#004, 008,  
H80  
7 mm  
Pulsed-RF  
83622B  
83623L  
3.5 mm (M)  
S-parameter  
Mixer-based  
#001, 004, 008  
[Option H50:  
83650B  
#004, 008  
(0.045 to 50)  
#001, 004, 008]  
85105A4  
MM-wave  
Controller  
33 to 110  
(waveguide  
bands)  
S-parameter  
Mixer-based  
(waveguide  
bands)  
83621B  
83621B  
(0.045 to 20)  
WR-22 (33 to 50)  
WR-19 (40 to 60)  
WR-15 (50 to 75)  
WR-10 (75 to 110)  
(0.045 to 20)  
[Option 050:  
83651B  
(0.045 to 50)]  
8510XF  
MM-wave  
Subsystem  
0.045 to 110  
(coaxial)  
S-parameter  
Mixer-based  
(ultra-  
83651B  
(0.045 to 50)  
83621B  
(0.045 to 20)  
1 mm (M)  
broadband)  
1. Although general purpose 8360 series synthesized sweepers (836x0B) can be used in place of the 8510-dedicated  
8360 series synthesized sweepers (836x1B), the following options are typically recommended: Option 004 (rear  
panel connectors) and Option 008 (1-Hz frequency resolution). These options are standard in the 8510-dedicated  
8360 series synthesized sweepers.  
Mixed sources: While mixing synthesized series is acceptable in multiple-source applications, the following areas  
must be considered:  
• RF source = 8340, LO source = 8340, system performance will be degraded substantially.  
• RF source = 8340, LO source = 8360, better system performance  
• RF source = 8360, LO source = 8360, faster step frequency measurements. Using the 8340 as either the RF  
source or the LO source will more than double the measurement time.  
2. All coaxial test port connectors are ruggedized connectors.  
3. These test sets provide access to four samplers directly.  
4. The following test set modules are available. Two test set modules must be ordered for complete waveguide  
S-parameter test set operation for each waveguide band:  
• Q85104A test set module (33 GHz to 50 GHz)  
• U85104A test set module (40 GHz to 60 GHz)  
• V85104A test set module (50 GHz to 75 GHz)  
• W85104A test set module (75 GHz to 110 GHz)  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
Sources  
The RF or microwave signal source in a VNA measurement system provides  
the incident signal used to stimulate the device-under-test (DUT). The DUT  
responds by reflecting part of the incident and transmitting the remaining  
part. By sweeping the frequency of the source, the frequency response of the  
DUT can be determined.  
There are two types of sources available; the General Purpose (GP) 8360 series  
synthesized sweepers (836x0B) and the 8510-dedicated 8360 series synthesized  
sweepers (836x1B). Note: All 8360A synthesizers have been replaced by 8360B  
synthesizers.  
Dedicated sources, such as the 836x1B shown in Table 4, are optimized  
for use in the 8510 network analyzer systems. They are configured without  
modulation capabilities or front panel keyboards. These sources include a  
1 Hz frequency resolution and rear-panel output connectors.  
Table 4. 8510-dedicated sources  
Product1  
83621B  
83631B  
83651B  
Frequency range  
45 MHz to 20 GHz  
45 MHz to 26.5 GHz  
45 MHz to 50 GHz  
For applications that require modulation capabilities, such as pulsed-RF  
measurements, the required source is a general purpose 836x0B synthesized  
sweeper, shown in Table 5. Order Option 004 to obtain rear panel connectors,  
and Option 008 for 1 Hz frequency resolution.  
Table 5. 8360 general purpose synthesizers  
Product1  
83620B  
83622B  
83623B  
83624B  
83630B  
83640B  
83650B  
Frequency range  
10 MHz to 20 GHz  
2 GHz to 20 GHz  
10 MHz to 20 GHz (high power)  
2 GHz to 20 GHz (high power)  
10 MHz to 26.5 GHz  
10 MHz to 40 GHz  
10 MHz to 50 GHz  
1. Although general purpose 8360 series synthesized sweepers (836x0B) can be used in place of the 8510-dedicated  
8360 series synthesized sweepers (836x1B), the following options are typically recommended: Option 004 (rear  
panel connectors) and Option 008 (1-Hz frequency resolution). These options are standard in the 8510-dedicated  
8360 series synthesized sweepers.  
Mixed sources: While mixing synthesized series is acceptable in multiple-source applications, the following areas  
must be considered:  
• RF source = 8340, LO source = 8340, system performance will be degraded substantially.  
• RF source = 8340, LO source = 8360, better system performance  
• RF source = 8360, LO source = 8360, faster step frequency measurements. Using the 8340 as either the  
RF source or the LO source will more than double the measurement time.  
8
Download from Www.Somanuals.com. All Manuals Search And Download.  
Upgrade considerations  
There are numerous reasons for upgrading the Agilent 8510 system. The 8510  
network analyzer maintains excellent performance while adapting to different  
measurement requirements such as basic component testing, on-wafer probing,  
pulsed device characterization, antenna and RCS (Radar Cross Section) meas-  
urements. Depending on the application, the 8510 network analyzer system  
can be reconfigured to meet other system measurement needs.  
Upgrading individual components  
Upgrading the 8510  
Figure 6 below illustrates how the 8510A or 8510B can be upgraded to an  
8510C.  
85103E  
8510A  
8510C  
85103F  
8510B  
Figure 6. 8510 upgrade paths  
Agilent 85103E Upgrades the 8510A to an 8510C by replacing the top unit and  
modifying the bottom unit of the 8510A. It includes on-site installation by an  
Agilent Customer Engineer (where available).  
Option 001 adds a rack modification kit for mounting in an 85043A  
system rack  
Option 002 adds an 8360 series source compatibility kit for 8517A/B test sets  
1
Option 003 adds an 8360 series source compatibility kit for 8514/15 and  
1
85110A test sets  
Option 004 adds an 85110L source compatibility kit  
Option 010 adds time domain to an 8510A with time domain previously  
installed  
Agilent 85103F Upgrades the 8510B to an 8510C by replacing the top unit on the  
8510B. It includes on-site installation by Customer Engineer (where available).  
Option 001 adds a rack modification kit for mounting in an 85043A system  
rack  
1
Option 002 adds an 8360 series source compatibility kit for 8517A/B test sets  
Option 003 adds an 8360 series source compatibility kit for 8514/15 and  
1
85110A test sets  
Option 004 adds an 85110L source compatibility kit  
Option 010 adds time domain to an 8510B with time domain previously installed  
1. Please review order instructions for serial number break.  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
Upgrading the 8510 to the 8530 microwave receiver for  
antenna measurements  
Figure 7 below illustrates how the 8510C can be upgraded to either an 8530A  
or 8530A and 8510C. For example, to upgrade an 8510C to an 8530A, order  
85395C with Option 111.  
85395C2  
85395C2  
Opt.111  
8510C  
8530A  
microwave receiver  
8530A  
microwave  
receiver  
85396A  
and  
8510C  
network analyzer  
capability  
Figure 7. 8510 to 8530 upgrade paths  
Agilent 85395C2 Upgrades any 8510C to an 8530A. Retains network analyzer  
capability. Includes on-site installation by Customer Engineer (where  
1
available). Other requirements may apply (such as 8530 firmware).  
The following options are available:  
Option 010 adds time domain capability  
Option 111 deletes network analyzer capability  
Agilent 85396A Adds 8510C Network Analyzer capability to any 8530A  
(equivalent to 8530A Option 011).  
Option 010 adds time domain capability  
1. For additional information, please refer to the Agilent 85395A/B/C and 85396A Upgrade Kits,  
literature number 5091-0948E.  
2. Requires an 85101C with serial prefix 3936A or lower  
10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Test set upgrades  
A single 8510 network analyzer can be configured to control up to four test  
sets. The operator can switch between test sets, without reconnections, using  
front panel controls. This arrangement is often referred to as the multiple  
test set configuration. Each test set must be equipped with Option 001 for IF  
switching. Other requirements may apply for RF/LO switching.  
For more details, please refer to product note 8510-14, Using Multiple Test  
Sets with the Agilent 8510C (literature number 5967-5886E).  
Agilent 8511A K01 or 08511-60008 Retrofits IF switching (Option 001) to any 8510  
test set for multiple test set operations.  
Expanding measurement capabilities  
Adding pulse capabilities (Option 008)  
To perform pulse measurements, the 8510 network analyzer must be retrofitted  
with Option 008. The measurement system must include either the 85110L  
(0.045 to 2 GHz) or 85110A (2 to 20 GHz) pulsed-RF S-parameter test set and  
two synthesizers (high power synthesizers may apply).  
• The 85110L requires an 83620B synthesizer with  
Options 001/004/008/H80 (for RF) and an 83620B with  
Options 004/008/H80 (for LO)  
• The 85110A requires an 83622B synthesizer with  
Options 001/004/008 (for RF) and an 83623L with  
Options 004/008 (for LO)  
Agilent 85111B Adds pulsed-RF measurement capability (Option 008) to the 8510C  
by adding new circuitry and includes on-site installation by an Customer  
Engineer (where available). To perform pulsed-RF measurements, the 8510  
with Option 008 must be used with either the 85110L or 85110A Pulsed-RF  
S-parameter test set. Other requirements may apply.  
Adding time domain capabilities  
With the time domain option, data from transmission or reflection measure-  
ments are converted from the frequency domain to the time domain via the  
inverse Fourier transform. The time domain data are presented on the CRT  
display showing the measured parameter value versus time.  
Agilent 85012C Adds time domain (Option 010) to an 8510C (customer installed).  
11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Upgrading systems  
Figure 8 shows the different 8510 systems and their components. This  
figure is an excellent guide when upgrading systems.  
Q-band  
U-band  
85106D #001  
0.045 to 110 GHz  
8510C  
85105A #050  
8517B  
83621B  
83651B  
Q-band  
U-band  
V-band  
W-band  
85106D  
OR  
OR  
OR  
OR  
OR  
OR  
33 to 110 GHz  
Waveguide  
systems  
+
+
8510C  
85105A  
83621B  
83621B  
V-band  
W-band  
8510XF  
85107B  
8510E  
8510SX  
Single-sweep  
high  
performance  
systems  
0.045 to 110 GHz  
0.045 to 50 GHz  
0.045 to 20 GHz  
8510C  
0.045 to 26.5 GHz  
8510C  
MM-wave subsystem  
83651B  
8510C  
8517B  
83651B  
8510C  
8515A  
83631B  
83621B  
83621B  
85108L  
85108A*  
85108A (50 GHz)  
0.045 to 2 GHz  
2 to 20 GHz  
2 to 50 GHz  
8510C  
Pulsed-RF  
or CW  
systems  
8510C  
85110L  
83620B #H80  
83620B #H80  
8510C  
85110A  
83622B  
83623L  
85110A (50 GHz)  
83650B  
83621B  
* 85108A also available in frequency range of 45 MHz to 20 GHz  
Figure 8. 8510 family of system solutions  
To upgrade from one system to another, simply order the pieces that are needed  
in the new system that are currently not in the existing system, an example  
is shown in Figure 9. Complete waveguide S-parameter systems require two  
1
test set modules for each band. Necessary accessories such as calibration kits,  
2
verification kits, test port cables or adapters must be ordered. See pages 17  
through 19 for more information.  
1. The following test set modules are available. Two test set modules must be ordered for complete waveguide  
S-parameter test set operation for each waveguide band:  
• Q85104A test set module (33 GHz to 50 GHz)  
• U85104A test set module (40 GHz to 60 GHz)  
• V85104A test set module (50 GHz to 75 GHz)  
• W85104A test set module (75 GHz to 110 GHz)  
2. Please refer to the RF & Microwave Test Accessories Catalog (literature number 5964-9527E) on detailed  
information on available adapters.  
12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Here is a checklist of items to consider when upgrading systems:  
1. Which 8510 do I have? The 8510C is recommended for the latest available  
measurement capabilities.  
• To upgrade an 8510A to an 8510C, order 85103E.  
• To upgrade an 8510B to an 8510C, order 85103F.  
• Make sure the appropriate options are ordered. For instance, order  
Option 010 if the 8510 has previously installed time domain.  
• Do I need pulse capabilities? If yes, order 85111B for 8510C.  
2. What test set do I need?  
• Refer to Table 3 for test set selection. Frequency range and application  
type are critical when selecting the appropriate test set.  
3. What source do I need?  
• Refer to Table 3 for the source(s) that’s required for the test set of choice.  
4. What measurement accessories do I need?  
• The appropriate measurement accessories will be determined mainly  
by the connector type of the device-under-test (DUT). These accessories  
include calibration (mechanical or electronic), verification kits, test port  
cables and/or adapters as appropriate. Refer to pages 17 through 19 for  
available products.  
System upgrades  
Upgrades available for existing 8510 systems to 8510XF single-sweep systems  
85106C w/ Opt.001 & 002  
85106D w/ Opt.001  
85109C w/ Opt.002  
85106C  
85106C w/ Opt.002  
85106D  
85107B  
85109C  
Upgrades  
from ...  
E7346A  
E7356A  
... to 85 GHz  
... to 110 GHz  
E7345A  
E7355A  
E7347A  
E7357A  
8510C  
Millimeter  
subsystem  
83651B  
83621B  
Customer  
owned  
equipment  
13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Upgrades for 85107B, 85109C  
Upgrade consists of two test heads, a millimeter test set controller, an 83621B  
for LO source, and rack. It does not include calibration kits or test port cables.  
E7345A upgrade to an 8510XF 85 GHz system  
E7355A upgrade to an 8510XF 110 GHz system  
The following options are available for both upgrades:  
Option 005 add 45 MHz to 2 GHz low frequency extension  
Option 006 add RF pass thru (provides coupled output of 50 GHz  
source for additional test sets. Additional test set(s) must have Option  
001 installed.)  
Upgrades for 85106C, 85106C with Option 002 (replaced 8350B/83540A with  
83621A/B), 85106D  
Upgrade consists of two test heads, a millimeter test set controller and an  
83651B for RF source. It does not include calibration kits, test port cables  
or rack.  
E7346A upgrade to an 8510XF 85 GHz system  
E7356A upgrade to an 8510XF 110 GHz system  
The following options are available for both upgrades:  
Option 005 add 45 MHz to 2 GHz low frequency extension  
Option 006 add RF pass thru (provides coupled output of 50 GHz  
source for additional test sets. Additional test set(s) must have Option  
001 installed.)  
Upgrades for 85106C with Options 001 and 002 (added 8517B, replaced 83621A/B  
with 83651A/B, and replaced 8350B/83540A with 83621A/B),  
85106D with Option 001 (added 8517B and replaced 83621B with 83651B),  
85109C with Option 002 (replaced 8350B/83540A with 83621A/B)  
Upgrade consists of two test heads and a millimeter test set. It does not  
include calibration kits, test port cables or rack.  
E7347A upgrade to an 8510XF 85 GHz system  
E7357A upgrade to an 8510XF 110 GHz system  
The following options are available for both upgrades:  
Option 005 add 45 MHz to 2 GHz low frequency extension  
Option 006 add RF pass thru (provides coupled output of 50 GHz  
source for additional test sets. Additional test set(s) must have Option  
001 installed.)  
14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Examples of upgrade paths  
85107B  
(45 MHz to 50 GHz)  
current system  
Upgrade to 85106D waveguide  
system to operate in V-band  
New equipment needed:  
85105A1 with  
Option 0502 millimeter-  
wave test set controller  
8510C  
8510C  
83621B synthesizer  
(LO source)  
V85104A 3 test set  
modules (quantity of 2)  
V11644A 3  
calibration kit  
V11645A 3  
85105A #050  
8517B  
8517B  
83651B  
verification kit  
83651B  
83621B  
Figure 10. Upgrading an 85107B 50 GHz system to an 85106D waveguide system to  
operate in V-band (or W-band)  
Measurement accessories  
There are measurement accessories for seven device connector types: 7 mm,  
3.5 mm, 2.92 mm, 2.4 mm, 1.85 mm, 1 mm and Type-N. Calibration kits include  
standards that are required for vector error correction. Verification kits  
include standards used to verify system performance specifications. Test port  
return cables extend the ports of the test set and connect to the device under  
test. Agilent 85130X adapter sets convert test set ports to the same connector  
type (acting as a test port saver) or to a different connector type.  
1. Standard 85105A already includes IF switching (Option 001)  
2. Option 050 provides a 50 GHz RF source switch. This is necessary in order to operate with the  
existing 83651B 50 GHz source and 8517B 50 GHz test set.  
3. To operate in W-band, replace all V-band components with W-band.  
15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Calibration kits  
Error correction requires that the systematic errors in the measurement  
system be characterized by measuring known devices (standards) over the  
frequency range of interest with the process of calibration. All calibration kits  
contain standards used for this purpose. The standards in the 3.5 mm, 2.4 mm  
and Type-N calibration kits use the precision slotless connector (PSC-3.5,  
PSC-2.4 and PSC-N). Unless otherwise noted all coaxial calibration kits include  
connector gauges and a torque wrench.  
Mechanical  
calibration  
kit  
Connector  
type  
Frequency  
range  
(GHz)  
Description  
85050B  
85050C  
7 mm  
7 mm  
0.045 - 18  
0.045 - 18  
Contains open and short circuits, fixed and sliding terminations.  
Precision kit. Contains standards for TRL calibration, including precision airline.  
Also contains open and short circuits and fixed termination.  
85050D  
85052B  
85052C  
7 mm  
0.045 - 18  
Economy kit. Contains open and short circuits and precision fixed termination.  
Gauges not included.  
3.5 mm  
3.5 mm  
0.045 - 26.5  
0.045 - 26.5  
Contains open and short circuits, fixed and sliding terminations and in-series  
adapters.  
Precision kit. Contains standards for TRL calibration, including precision airlines.  
Also contains open and short circuits, fixed terminations and in-series adapters.  
Gauges not included.  
85052D  
85054B  
85054D  
3.5 mm  
Type-N  
Type-N  
0.045 - 26.5  
0.045 - 18  
0.045 - 18  
Economy kit. Contains open and short circuits, precision fixed termination, and in-series  
adapters. Gauges not included.  
Contains open and short circuits, fixed and sliding terminations, in-series adapters  
and 7 mm-to-Type-N adapters.  
Economy kit. Contains open and short circuits, fixed terminations, in-series  
adapters and 7 mm-to-Type-N adapters. Gauges not included.  
85056A  
85056D  
2.4 mm  
2.4 mm  
0.045 - 50  
0.045 - 50  
Contains open and short circuits, fixed and sliding terminations and in-series adapters.  
Economy kit. Contains open and short circuits, fixed terminations and in-series  
adapters. Gauges not included.  
85056K  
85059A  
2.92/2.4 mm  
1 mm  
0.045 - 50  
Contains 2.4 mm open and short circuits, fixed loads and 2.92 mm adapters.  
0.045 - 110  
Broadband coaxial precision calibration kit consists of 1 mm shorts, opens, fixed  
loads and in-series adapters. It also includes offset-shorts covering 50 to 110 GHz.  
Gauges not included.  
11904S  
2.92 mm  
0.045 - 40  
Must be used with Agilent 85056A/D 2.4 mm calibration kit. Includes four 2.92 mm-to-  
2.4 mm adapters. Gauges not included.  
X11644A  
P11644A  
K11644A  
R11644A  
Q11644A  
U11644A  
V11644A  
W11644A  
WR-90  
WR-62  
WR-42  
WR-28  
WR-22  
WR-19  
WR-15  
WR-10  
8.2 - 12.4  
12.4 - 18  
18 - 26.5  
26.5 - 40  
33 - 50  
Contains standards for TRL calibration. Includes precision waveguide  
section, short circuit and fixed or sliding terminations. Gauges not included.  
40 - 60  
50 - 75  
75 - 110  
16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Verification kits  
Verification kits are used to verify the performance specifications of an 8510  
system. All kits include a precision Zo airline, mismatched airline and fixed  
attenuators. Measured data and uncertainties traceable to the U.S. National  
Institute of Standards and Technology (NIST) are included with each kit.  
Compliance with MIL-STD 45662A is available for an extra charge  
Choose a verification kit for each connector type required.  
Verification  
kit  
Connector  
type  
Frequency  
range (GHz)  
Verification Connector  
Frequency  
range (GHz)  
kit  
type  
85051B  
85053B  
85055A  
85057B  
7 mm  
0.045 - 18  
0.045 - 26.5  
0.045 - 18  
0.045 - 50  
R11645A  
Q11645A  
U11645A  
V11645A  
W11645A  
WR-28  
WR-22  
WR-19  
WR-15  
WR-10  
26.5 - 40  
33 - 50  
40 - 60  
50 - 75  
75 - 110  
3.5 mm  
Type-N  
2.4 mm  
Test port cables and adapters  
Test port cables and adapter sets are available for various connector types.  
Special test port adapter sets convert the rugged ports of the network analyzer  
test set to the desired connector interface. Each kit contains two adapters, one  
male and one female. Both the cables and the test port adapters have one spe-  
cial female connector which is designed to connect directly to the 3.5 mm test  
port (2.4 mm for 8517B). This side of the cable or adapter can only be connected  
to the test set port and cannot be mated to a standard 3.5 mm (or 2.4 mm)  
male connector. Choose one of the configurations shown.  
Configuration A. This cable arrangement is for applications where the device  
under test is connected directly to the test set port. This setup offers the best  
mechanical rigidity for device connection. To adapt the test set port (port 1) to  
the device under test, choose the appropriate special adapter set. In addition  
to converting the test port to the desired interface, these adapters also func-  
tion as “test port savers” which protect the test set from damage and wear due  
to heavy use.  
Configuration A  
Network analyzer  
test set  
Test port  
adapter  
Device  
under  
test  
For Agilent 8514B/8515A/85110A test sets (3.5 mm rugged test port connectors)  
Cables/adapter  
Connector type  
(on device side of cable/adapter)  
For 3.5 mm devices  
For 7 mm devices  
For Type-N devices  
85131C semi-rigid cable or  
85131E flexible cable  
85130D adapter set  
3.5 mm (f)  
3.5 mm (f )  
3.5 mm (m and f)  
Single cable  
85132C semi-rigid cable or  
85132E flexible cable  
85130B adapter set  
7 mm  
7 mm  
7 mm  
Use 7 mm cables and the 7 mm-to-Type-N  
adapters included in the 85054B/D  
Type-N calibration kit  
For Agilent 8517B test sets (2.4-mm rugged test port connectors)  
Cables/adapter  
Connector type  
(on device side of cable/adapter)  
For 2.4 mm devices  
For 3.5 mm devices  
For 7 mm devices  
85133C semi-rigid cable or  
85133E flexible cable  
85130G adapter set  
2.4 mm (f)  
2.4 mm (f )  
2.4 mm (m and f)  
85134C semi-rigid cable or  
85134E flexible cable  
85130F adapter set  
3.5 mm (f)  
3.5 mm (f )  
3.5 mm (m and f)  
85135C semi-rigid cable or  
85135E flexible cable  
85130E adapter set  
7 mm  
7 mm  
7 mm  
17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Configuration B. This cable arrangement is for applications where the device  
under test is connected between cable ends. This setup offers more flexibility  
when connecting to the device under test.  
Configuration B  
Network analyzer  
test set  
For 8514B/8515A/85110A test sets (3.5 mm rugged test port connectors)  
Cables/adapter  
Connector type  
(on device side of cable/adapter)  
Cable set  
For 3.5 mm devices  
For 7 mm devices  
For Type-N devices  
85131D semi-rigid cable set or  
85131F flexible cable set  
3.5 mm (m and f)  
3.5 mm (m and f)  
85132D semi-rigid cable set or  
85132F flexible cable set  
7 mm  
7 mm  
Device  
under  
test  
Use 7 mm cables and the 7 mm-to-Type-N  
adapters included in the 85054B/D  
Type-N calibration kit  
For 85110L Test sets (7 mm rugged test port connectors)  
Cables/adapter  
Connector type  
(on device side of cable/adapter)  
For 7 mm devices  
11857D Cable pair  
7 mm  
For 8517B Test sets (2.4 mm rugged test port connectors)  
Cables/adapters  
Connector type  
(on device side of cable/adapter)  
For 2.4 mm devices  
For 3.5 mm devices  
For 7 mm devices  
85133D semi-rigid cable set or  
85133F flexible cable set  
2.4 mm (m and f)  
2.4 mm (m and f)  
85134D semi-rigid cable set or  
85134F flexible cable set  
3.5 mm (m and f)  
3.5 mm (m and f)  
85135D semi-rigid cable set or  
85135F flexible cable set  
7 mm  
7 mm  
For 8510XF systems (1 mm test port connectors)  
Cables/adapters  
Connector type  
For 1 mm devices  
11500I (8.8 cm) test port cable  
11500J (16 cm) test port cable  
11500K (20 cm) test port cable  
11500L (24 cm) test port cable  
1 mm (f and f)  
1 mm (m and f)  
1 mm (m and f)  
1 mm (m and f)  
For V-band waveguide devices  
For W-band waveguide devices  
V281C adapter  
V281D adapter  
1 mm (f) to V-band  
waveguide devices  
1 mm (m) to V-band  
waveguide devices  
W281C adapter  
W281D adapter  
1 mm (f) to W-band  
waveguide devices  
1 mm (m) to W-band  
waveguide devices  
18  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Related literature  
Pub. number  
5091-8970E  
5965-8837E  
5091-8967E  
5091-8484E  
5091-8969E  
Agilent 8510 Network Analyzer, Color Brochure  
Agilent 8510 System Solutions  
Agilent 8510 Family Network Analyzer, Configuration Guide  
Agilent 8510 Family Network Analyzer, Data Sheet  
Agilent 85103C/D Upgrade Packages to the 8510C  
Agilent 85395A/B/C and 85396A Upgrade Kits,  
(8510 to 8530 upgrades)  
5091-0948E  
5964-6162E  
Agilent 8360B Series Synthesized Swept Signal Generators,  
Data Sheet  
Product overviews  
Agilent 85106D Millimeter-wave Network Analyzer System  
Agilent 85108A/L CW/Pulsed Network Analyzer Systems  
Agilent 8510XF 110 GHz Single-Sweep Systems  
Agilent 85060 Series and 85090 Series Electronic  
Calibration Modules and PC Interface  
5964-4229E  
5091-8965E  
5965-9888E  
5963-3743E  
Product notes  
Agilent PN 8510-14, Using Multiple Test Sets  
with the 8510C  
5967-5886E  
19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Agilent Technologies’ Test and  
Measurement Support, Services,  
and Assistance  
By internet, phone, or fax, get assistance  
with all your test and measurement needs.  
For more information on the 8510  
VNA visit:  
Agilent Technologies aims to maximize  
the value you receive, while minimizing  
your risk and problems. We strive to  
ensure that you get the test and measure-  
ment capabilities you paid for and obtain  
the support you need. Our extensive sup-  
port resources and services can help you  
choose the right Agilent products for your  
applications and apply them successfully.  
Every instrument and system we sell has  
a global warranty. Support is available for  
at least five years beyond the production  
life of the product. Two concepts underlie  
Agilent's overall support policy: "Our  
Promise" and "Your Advantage."  
Online assistance:  
Phone or Fax  
United States:  
(tel) 800 452 4844  
Get the latest information on the products  
and applications you select.  
Canada:  
(tel) 877 894 4414  
(fax) 905 282 6495  
China:  
(tel) 800 810 0189  
(fax) 800 820 2816  
Our Promise  
Europe:  
Our Promise means your Agilent test and  
measurement equipment will meet its  
advertised performance and functionality.  
When you are choosing new equipment,  
we will help you with product information,  
including realistic performance specifica-  
tions and practical recommendations from  
experienced test engineers. When you  
use Agilent equipment, we can verify that  
it works properly, help with product opera-  
tion, and provide basic measurement  
assistance for the use of specified capa-  
bilities, at no extra cost upon request.  
Many self-help tools are available.  
(tel) (31 20) 547 2323  
(fax) (31 20) 547 2390  
Japan:  
(tel) (81) 426 56 7832  
(fax) (81) 426 56 7840  
Korea:  
(tel) (82 2) 2004 5004  
(fax) (82 2) 2004 5115  
Latin America:  
(tel) (305) 269 7500  
(fax) (305) 269 7599  
Your Advantage  
Taiwan:  
Your Advantage means that Agilent offers  
a wide range of additional expert test and  
measurement services, which you can  
purchase according to your unique techni-  
cal and business needs. Solve problems  
efficiently and gain a competitive edge by  
contracting with us for calibration, extra-  
cost upgrades, out-of-warranty repairs,  
and on-site education and training, as  
well as design, system integration, project  
management, and other professional engi-  
neering services. Experienced Agilent  
engineers and technicians worldwide can  
help you maximize your productivity, opti-  
mize the return on investment of your  
Agilent instruments and systems, and  
obtain dependable measurement accuracy  
for the life of those products.  
(tel) 0800 047 866  
(fax) 0800 286 331  
Other Asia Pacific Countries:  
(tel) (65) 6375 8100  
(fax) (65) 6836 0252  
Product specifications and descriptions  
in this document subject to change  
without notice.  
Copyright © 2002  
Agilent Technologies, Inc.  
Printed in USA, June 13, 2002  
5968-6423E  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Acer Laptop 2400 Series User Manual
Acer Projector P5271 Series User Manual
Acnodes Server KD 6176 User Manual
AEG Ventilation Hood DL 7275 User Manual
Alpine Car Amplifier MRX M100 User Manual
Altec Lansing Intercom System LEF 3L User Manual
Axion Oven 16 3903 User Manual
Axis Communications Security Camera M3011 User Manual
Baldor Music Mixer MN754 User Manual
Barco Crib Toy w18 User Manual