3M Barcode Reader 390G User Manual

3MAttestAuto-reader 390G  
1
2
3
4
5
5
6
7
8
9
0
Operator’s Manual  
Manuel de l’utilisateur  
Benutzerhandbuch  
Manuale dell’operatore  
Manual del usuario  
Manual del operador  
Bedieningshandleiding  
Användarhandbok  
Brugsvejledning  
-
-
=
H
F
S
M
A
:
I
T
Manual do Utilizador  
Manual do Operador  
Εγχειρίδιο Χειριστή  
Instrukcja obsługi  
Příručka pro obsluhu  
Priročnik za uporabo  
Инструкция по эксплуатации  
Upute za korištenje  
Operatör Kılavuzu  
操作指南  
Brukerhåndbok  
Käyttöopas  
อปฏงาน  
작업자 매뉴얼  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Warnings and Safety Precautions  
The following warnings and precautions should be followed to avoid unsafe actions that  
could result in personal injury or damage to the instrument.  
PREFACE  
Manual Revision History  
Date  
Revision Reason For Revision  
WARNING: To reduce the risk associated with hazardous voltage  
Use indoors only.  
November 2012  
A
Initial Issue  
To clarify, in the Indications for Use section, that  
Attest™ 1294 Biological Indicators are used to  
monitor the ethylene oxide sterilization process.  
February 2013  
July 2013  
B
Do not use the equipment if it is not working properly or if it has suffered any damage.  
Use only the power supply specified for this product and certified for the country  
of use.  
To revise: the Warranty statement; and the WEEE  
symbol and associated explanatory text.  
C
Content Disclaimers  
CAUTION: To reduce the risk of injury or instrument damage  
Pictorial Disclaimer  
Do not spill liquid onto or into the instrument. Do not immerse the unit in liquid.  
Always unplug the 3M™ Attest™ Auto‑reader 390G and allow to cool before cleaning.  
Clean external surfaces using only the instructions provided by the manufacturer.  
Do not open the instrument housing – there are no user serviceable parts.  
The instrument must be returned to the manufacturer for repair.  
Sample printouts, graphics, displays and screens are for information and illustration  
purposes only, and shall not be used for clinical or maintenance evaluations. Data shown  
in sample printouts and screens do not reflect actual patient names or test results.  
Hardware Disclaimer  
The 3M™ Attest™ Auto‑reader 390G hardware specifications are subject to change.  
The system images, hardware components and hardware specifications included in  
the manual may not match the system installed. Any changes or modifications to the  
authorized system installation have been verified as compatible with the functionality  
outlined in this document.  
Allow the 3M™ Attest™ Rapid Readout Biological Indicator to cool for the  
recommended time period before activating. Activating or excessive handling of the  
biological indicator before cooling may cause the glass ampoule to burst.  
Wear safety glasses when activating the 3M™ Attest™ Rapid Readout Biological  
Indicator.  
INDICATIONS FOR USE  
The 3M™ Attest™ Auto‑reader 390G is designed to incubate and automatically read the  
3M™ Attest™ Rapid Readout Biological Indicator for Ethylene Oxide, 1294, at 37°C for a  
final fluorescent result at 4 hours.  
WARNING: To reduce the risk associated with incorrect results  
WARRANTY  
Instrument is to be used by operators familiar with the device, device functionality, and  
Operator’s Manual.  
The 3M™ Attest™ Auto‑reader 390G has a one year limited product warranty. The US  
warranty, remedy and limitations are described in the Price Quote form and in the Direct  
Price Pages. For countries outside of the US, any warranty is established by the subsidiary  
with responsibility for service of the 3M™ Attest™ Auto‑reader 390G.  
Do not place the instrument in environment exposed to sunlight or strong  
incandescent light.  
Do not place instrument close to any device that emits a strong electro‑magnetic field.  
Do not use on a vibrating surface.  
Serial Number  
For easy identification, each 3M™ Attest™ Auto‑reader 390G has a unique serial number  
printed on a label found on the back of the unit and displayed on the left side of the upper  
Match the cap color of the 3M™ Attest™ Rapid Readout Biological Indicator with the  
color‑coded configuration sticker surrounding the 3M™ Attest™ Auto‑reader 390G  
incubation wells.  
row of text on the LCD Display when the user depresses the  
your serial number in this manual for future reference: ___________________.  
button. Please record  
Do not remove or change placement of 3M™ Attest™ Rapid Readout Biological  
Indicator once it is placed into a well.  
SAFETY AND PRECAUTIONS  
Do not remove the 3M™ Attest™ Rapid Readout Biological Indicator from the  
incubation well until the (+) or (‑) symbol on the LCD panel indicates the test  
is complete.  
The 3M™ Attest™ Auto‑reader 390G and its related devices and accessories are  
designed to provide safe and reliable service when used according to the instructions  
provided. Please read, understand, and follow all safety information contained in the  
instructions for use included with the 3M™ Attest™ Auto‑reader 390G and 1294 Rapid  
Readout Biological Indicator devices prior to use. Use this equipment only for the purpose  
described in this Operator’s Manual. Retain these instructions for future reference.  
Caution: To reduce the risk associated with incorrect results  
The unit is designed to be used only with the Power Supply module and Ethernet cable  
supplied by 3M. If this product is used in a manner not specified, the protection provided  
by the product may be impaired.  
To avoid the possibility of the 3M™ Attest™ Rapid Readout Biological Indicator  
vial absorbing fluorescent residue from a chemical indicator or tape, place the  
3M™ Attest™ Rapid Readout Biological Indicator vial so it does not come in direct  
contact with chemical indicators or tape.  
Explanation of Single Word Consequences  
Warning: Indicates a hazardous situation, which, if not avoided, could result in  
death or serious injury.  
Caution: Indicates a hazardous situation, which, if not avoided, could result in  
minor or moderate injury.  
2
Download from Www.Somanuals.com. All Manuals Search And Download.  
HEALTH AND SAFETY INFORMATION  
EXPLANATION OF PRODUCT AND PACKAGING  
LABEL SYMBOLS  
Device Safety Compliance  
Attention ‑ Refer to the instructions for use  
The 3M™ Attest™ Auto‑reader 390G complies with the following standards as  
demonstrated by the CB Scheme Certificate and test report issued by Underwriters  
Laboratories (UL):  
Waste Electrical and Electronic Equipment (WEEE) and EU Battery Directive.  
This symbol indicates that both the device and the lithium ion battery  
contained therein need to be disposed of properly.  
IEC 61010‑1 (2001) Second Edition. Safety requirements for electrical equipment for  
measurement, control, and laboratory use ‑ Part 1: General requirements  
IEC 61010‑2‑010 (2003) Second Edition. Safety requirements for electrical equipment  
for measurement, control, and laboratory use ‑ Part 2‑010: Particular requirements for  
laboratory equipment for the heating of materials  
Li  
UL Listed to US and Canadian Safety Standards  
Mark of Conformity to European Directives  
Direct current  
The 3M™ Attest™ Auto‑reader 390G is listed as Laboratory equipment and carries the  
UL mark with adjacent indicators “C” and “US” based on compliance to the standards UL  
61010‑1 and CAN/CSA 22.2 No. 61010‑1.  
Compliant to all applicable ACMA regulatory arrangements (RCM)  
Manufacturer  
The 3M™ Attest™ Auto‑reader 390G complies with the CE mark related to the Low  
Voltage Directive (LVD) 2006/95/EC as confirmed in the Declaration of Conformity.  
The 3M™ Attest™ Auto‑reader 390G complies with the RoHS Directive, Directive  
2011/65/EU of the European Parliament and of the Council of 8 June 2011 on  
the restriction of the use of certain hazardous substances in electrical and  
electronic equipment.  
Catalog number  
REF  
Authorized representative for the European Community  
The 3M™ Attest™ Auto‑reader 390G complies with the WEEE Directive, Directive  
2012/19/EU of the European Parliament and of the Council of 04 July 2012 on waste  
electrical and electronic equipment (WEEE).  
Serial number  
1294  
EMC Compliance  
The 3M™ Attest™ Auto‑reader 390G complies with the following EMC standards as  
confirmed in the Certificate of Compliance generated by 3M:  
The 3M™ Attest™ Auto‑reader 390G is not designed to incubate  
3M™ Attest™ 1264 EO Biological Indicators, which have a sterilant  
entry port in the center of the top of the cap. The 3M™ Attest™  
Auto‑reader 390G is designed to incubate and read only 3M™ Attest™  
Rapid Readout Biological Indicators for EO, catalog number 1294. Other  
biological indicators (BIs) are not compatible with this device and cannot  
be used.  
IEC 61326‑1 Electrical equipment for measurement, control and laboratory use ‑ EMC  
requirements ‑ Part 1: General requirements  
The 3M™ Attest™ Auto‑reader 390G complies with the EMC requirements of the CE  
mark EMC Directive 2004/108/EC.  
The 3M™ Attest™ Auto‑reader 390G complies with the Australian and New Zealand  
electrical safety and electromagnetic compatibility requirements as confirmed in the  
Supplier’s Declaration of Conformity that is linked to the Australian/New Zealand RCM  
(Regulatory Compliance Mark).  
Note: This equipment has been tested and found to comply with the limits for a Class A  
digital device pursuant to Part 15, Subpart B, of the FCC Rules. These limits are designed  
to provide a reasonable protection against harmful interference when the equipment is  
operated in a commercial environment. This equipment generates, and can radiate radio  
frequency energy and, if not installed and used in accordance with the instruction manual,  
may cause harmful interference to radio communications. Operation of this equipment  
in a residential area is likely to cause harmful interference in which case the user will be  
required to correct the interference at their own expense.  
This Class A digital device meets all requirements of the Canadian Interference‑Causing  
Equipment Regulations.  
Class 1 LED Product  
An investigation was conducted on the 3M™ Attest™ Auto‑reader 390G per the  
requirements of IEC 60825‑1 Ed. 1.2 (A2:2001). The 3M™ Attest™ Auto‑reader 390G is a  
Class 1 LED product with Class 1 Internal Radiation.  
The LED apertures are located at the bottom of each Incubator/Reader well perpendicular  
to the opening of each well.  
NOTE: Class 1 products are safe for unprotected exposure of the eyes.  
Caution ‑ Use of controls or adjustments or performance of procedures other than  
those specified herein may result in hazardous radiation exposure.  
Note: The use of the term “controls” in the above caution statement as required by  
IEC 60825‑1 Ed. 1.2 (A2:2001) refers to mechanisms by which the 3M™ Attest™  
Auto‑reader 390G may be controlled and not to the use of biological indicator controls.  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
SPECIFICATIONS  
Power Supply Specifications  
Power supply  
Operating Condition  
Units  
Voltage Range  
100‑240  
Volts AC  
Frequency  
Current  
50/60  
0.6  
Hertz  
Amps  
Environmental Operating Conditions  
Environmental Condition  
Operating Condition  
Units  
Altitude  
3000 (max)  
Meters  
Operating Temperature  
Storage Temperature  
Operating Relative Humidity  
Voltage Range  
16 – 29  
ºC  
‑10 to 50  
ºC  
20‑80 (non‑condensing)  
%
12  
Volts DC  
Frequency  
DC  
Current  
1.2  
Amps  
Installation/Over Voltage  
Pollution Degree  
Category II  
2
INSTRUCTIONS FOR USE – Initial Set‑up  
Choose Location  
WARNING:  
Place 3M™ Attest™ Auto‑reader 390G on a firm and level surface, away from sunlight  
and incandescent light. To save counter space, units may be stacked two high.  
Do not place the instrument in environment exposed to sunlight or strong  
incandescent light.  
Do not place instrument close to any device that emits a strong  
electro‑magnetic field.  
Do not use on a vibrating surface.  
Select & Install Correct Power Plug  
WARNING: Use only the power supply specified for this product and certified for  
the country of use.  
The 3M™ Attest™ Auto‑reader 390G power supply comes with multiple blade  
configurations. Install the blade configuration appropriate for your country into the power  
supply and plug into appropriate AC grounded outlet.  
Wait 30 Minutes Before Using  
Once plugged in, allow a 30 minute warm‑up period before placing 3M™ Attest™ Rapid Readout Biological Indicators into the incubation wells. The C1 caution code will disappear  
when proper incubation temperature is reached.  
Special Considerations  
It is recommended that the unit be left on to eliminate warm‑up periods. The unit may be turned off when it is anticipated that it will not be used for an extended period.  
4
Download from Www.Somanuals.com. All Manuals Search And Download.  
INSTRUCTIONS FOR USE – Wells, Buttons, & Display Interpretation  
Please familiarize yourself with the face of the 3M™ Attest™ Auto‑reader 390G:  
INSTRUCTIONS FOR USE – 3M™ Attest™  
Auto‑reader 390G Configuration  
Wells  
Note: Ideally, the 3M™ Attest™ Auto‑reader 390G should be empty (i.e., have no  
incubating BIs) when it is reconfigured.  
1. Incubation/Reader Wells – The 3M™ Attest™ Auto‑reader 390G has ten  
incubation wells.  
Buttons  
Press the  
and  
buttons on the front of the 3M™ Attest™ Auto‑reader 390G  
simultaneously for 3 seconds to enter configuration mode. Once in configuration mode,  
2. Information / Up Arrow Button  
use the Up and Down buttons to set the 3M™ Attest™ Auto‑reader 390G  
3. Temperature / Down Arrow Button  
clock and audible alarm as desired. The  
button acts as an “Enter” key and moves the  
flashing cursor to the next changeable setting.  
4. Buzzer Alarm OFF / Enter Button  
Setting the 3M™ Attest™ Auto‑reader 390G Clock  
Information Display  
The 3M™ Attest™ Auto‑reader 390G date and time are displayed in this format:  
YYYY‑MM‑DD HH:MM, with the time displayed using a 24 hour clock. To change the year,  
5. LCD Display – Display is backlit  
6. LCD Display – Example of Error Message for well #6  
7. LCD Display – Example of Remaining Incubation Time, in minutes, for well #10  
8. LCD Display – Example of Negative ( ‑ ) BI result for well #9  
9. LCD Display – Example of Positive ( + ) BI result for well #8  
10. LCD Display – Alarm Enabled Indicator  
if necessary, push the up or down buttons to set desired year, then press  
to advance  
to the month, adjust as necessary and press  
to advance to the day. Use the  
button to move to the time and set the hours and minutes.  
Setting the Audible Alarm Function  
Color‑Coded Configuration Indicator  
11. Incubation well configuration sticker is color‑coded green  
BI Crusher  
Move across the top row of text by pushing the  
button. The flashing “B” in Buzzer  
indicates this is a changeable setting. It is recommended this be set at “Buzzer Enabled”  
to audibly alert the operator to a positive BI but it may be set to “Buzzer Disabled” by  
using either the up or down button. Note: If Buzzer Disabled is selected, and the changes  
accepted, the speaker icon ( ) will disappear from the LCD panel and the alarm will be  
disabled for all positive 3M™ Attest™ Rapid Readout Biological Indicators, error codes,  
and caution code alarm warnings. Note: The 3M™ Attest™ Auto‑reader 390G display  
backlight flashes whenever an audible alarm sounds. Disabling the audible alarm does  
NOT disable the LCD display backlight flashing.  
12. BI Crusher Well  
Press  
to select the desired option and advance to the Accept screen.  
Accept Screen  
Use the up or down buttons to toggle between Yes and No, press  
to accept changes  
(Yes selected) or retain current settings (No selected) and exit the configuration mode. The  
LCD panel will return to the standard display.  
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
INSTRUCTIONS FOR USE ‑ Readout Procedure  
INSTRUCTIONS FOR USE – Additional Information  
Instructions for Biological Indicator (BI) Activation and Incubation  
BI Placement  
Remaining incubation time, in minutes, will be displayed on the LCD display to indicate  
the 3M™ Attest™ Rapid Readout Biological Indicator is incubating in the reader well and  
should not be touched. The time remaining display indicates the results are not available  
and incubation is in progress.  
CAUTION: To avoid the possibility of the 3M™ Attest™ Rapid Readout  
Biological Indicator vial absorbing fluorescent residue from a chemical indicator or  
tape, place the 3M™ Attest™ Rapid Readout Biological Indicator vial so it does not  
come in direct contact with chemical indicators or tape.  
Do not remove 3M™ Attest™ Rapid Readout Biological Indicator until the (+) or (‑)  
symbol on the LCD panel indicates the test is complete.  
Note: If 3M™ Attest™ Rapid Readout Biological Indicator is inadvertently removed  
before the incubation period is complete, an alarm will sound and a C2 Caution code will  
be displayed on the LCD panel under that well. You must return the 3M™ Attest™  
Rapid Readout Biological Indicator to the reader well within 10 seconds or the  
results will be lost.  
WARNING: Match the green cap color of the 3M™ Attest™ 1294 Rapid  
Readout Biological Indicator with the green color‑coded configuration sticker  
surrounding the 3M™ Attest™ Auto‑reader 390G incubation wells.  
CAUTION: Wear safety glasses when activating 3M™ Attest™ Rapid Readout  
Biological Indicators.  
WARNING: Do not remove or change placement of 3M™ Attest™ Rapid  
Readout Biological Indicator once it is placed into a well. Doing so may result in loss of  
results and the 3M™ Attest™ Rapid Readout Biological Indicator test may be invalid.  
Follow this procedure for every 3M™ Attest™ 1294 Rapid Readout Biological Indicator to  
be read.  
While wearing safety glasses, CLOSE the cap of the  
cooled 3M™ Attest™ 1294 Rapid Readout Biological  
Indicator (BI) by pushing cap down.  
INSTRUCTIONS FOR USE – Positive Control  
Positive Control  
Activate and incubate at least one non‑processed 3M™ Attest™ 1294 Rapid  
Readout Biological Indicator to use as the positive control each day that a processed  
3M™ Attest™ 1294 Rapid Readout Biological Indicator is incubated and read by the unit.  
Write a “C” (for “control”) and the date on the label. The positive control can be placed in  
any incubation well. The positive control should be the same part number and have the  
same lot number as the processed 3M™ Attest™ Rapid Readout Biological Indicator in  
the unit.  
CRUSH the glass media ampoule in the Crusher Well  
built into the top of the 390G Auto‑reader.  
The positive control biological indicator helps confirm:  
correct incubation temperatures are met (heater block reading of 37ºC +/‑2ºC);  
viability of spores has not been altered due to improper storage temperature, humidity,  
or proximity to chemicals;  
capability of media to promote rapid growth; and  
proper functioning of 3M™ Attest™ Auto‑reader 390G components.  
Holding BI by the cap, TAP the bottom of the vial on a  
tabletop until media wets spore strip at bottom of vial.  
Do not tap BI on Auto‑reader.  
INSTRUCTIONS FOR USE – Interpretation of Results  
Fluorescent Results  
After placing a 3M™ Attest™ Rapid Readout Biological Indicator into an incubation  
well, the 3M™ Attest™ Auto‑reader 390G LCD panel will display ***, **, * and then  
the remaining minutes of incubation directly below the incubating BI to indicate that  
incubation/reading is in progress.  
PLACE the crushed BI in color‑coded incubation well.  
Observe 3M™ Attest™ Auto‑reader 390G LCD panel  
to confirm correct placement of the BI. If the BI is  
correctly seated, the 3M™ Attest™ Auto‑reader 390G  
LCD panel will display ***, **, * and then 240 remaining  
minutes of incubation directly below the incubating BI.  
A (+) will display on the LCD panel directly under the BI and an alarm will sound (if the  
sound feature is enabled) as soon as a positive 3M™ Attest™ Rapid Readout Biological  
Indicator result is detected.  
At the end of the four hour incubation time, if a negative 3M™ Attest™ Rapid Readout  
Biological Indicator result is detected, a (‑) will display on the LCD panel directly under  
the incubating BI. The final negative fluorescent result (‑) for a processed 3M™ Attest™  
1294 Rapid Readout Biological Indicator after 4 hours of incubation indicates an  
acceptable sterilization process. The positive control should read positive (+).  
RECORD BI result.  
After either the + or – symbol is displayed, and the result is recorded, the 3M™ Attest™  
Rapid Readout Biological Indicator can be removed from the well. The 3M™ Attest™  
Rapid Readout Biological Indicator should not be removed or touched when the display  
indicates remaining incubation time.  
WARNING: Do not remove the 3M™ Attest™ 1294 Rapid Readout Biological  
Indicator from the incubation well until the (+) or (‑) symbol on the LCD panel indicates  
the test is complete.  
The positive (unprocessed) control 3M™ Attest™ 1294 Rapid Readout Biological Indicator  
must provide a positive fluorescent result (+). Processed 3M™ Attest™ Rapid Readout  
Biological Indicator results are not valid until the positive control reads fluorescent positive  
(+). With a processed 3M™ Attest™ Rapid Readout Biological Indicator, a positive (+)  
result indicates a sterilization process failure.  
Act on any positive test results immediately. Determine the cause of the positive  
3M™ Attest™ Rapid Readout Biological Indicator and retest the sterilizer following  
current facility policies and procedures.  
6
Download from Www.Somanuals.com. All Manuals Search And Download.  
Optional Visual pH Color Change Result  
Easy Configuration  
3M™ Attest™ 1294 Rapid Readout Biological Indicators may be further incubated for  
a visual pH color change. Due to the high sensitivity of the fluorescent result, there is no  
advantage to incubating the 3M™ Attest™ 1294 Rapid Readout Biological Indicators  
beyond 4 hours. This optional further incubation is not intended for routine use but may  
be used for special studies. A humidified incubator will be required to avoid media dry out.  
In the case of a positive control BI, a yellow color change of the growth media will appear  
within 7 days. Any observation of a yellow color within the vial indicates a positive result.  
The pH color change result can be recorded and the BI discarded as soon as the yellow  
color is observed and the result recorded. In the case of a processed BI, a media color  
change from blue‑green to yellow indicates a sterilization process failure. A negative pH  
color change result, i.e., media remains blue‑green, can be assessed at 7 days.  
The 3M™ Attest™ Auto‑reader 390G may be configured using either the buttons located  
on the front of the unit or using the optional 3M™ Attest™ Auto‑reader 390G Web App.  
Optional 3M™ Attest™ Auto‑reader 390G Web App  
If desired, the 3M™ Attest™ Auto‑reader 390G may be connected to a PC via an Ethernet  
connection for additional capabilities:  
Remotely view current status of BIs in the 3M™ Attest™ Auto‑reader 390G via PC  
or mobile hand held devices. For full functionality, the use of an HTML5 compliant  
browser with JavaScript enabled is recommended.  
Remotely view and configure 3M™ Attest™ Auto‑reader 390G settings.  
View the last 100 BI records.  
Print out documentation using a local or networked printer and/or save documentation  
as an electronic file.*  
INSTRUCTIONS FOR USE – Disposal  
Instrument contains a lithium battery. Dispose of instrument in accordance with federal,  
state, and local requirements.  
Note: To access the 3M™ Attest™ Auto‑reader 390G Web App within a secure network,  
the PC or mobile device must have access to the secure network; options include  
connecting via a network Ethernet cable, signing in to the facility’s wireless network, or  
connecting via a virtual private network (VPN).  
FEATURES OF THE 3M™ ATTEST™ AUTO‑READER 390G  
* Refer to your facility’s record keeping policy and procedure.  
Audible Alarm  
An audible alarm will sound, the LCD display backlight will flash, and a (+) symbol will  
be displayed on the LCD panel directly under the incubating BI when a positive result is  
detected. An alarm will also sound if a 3M™ Attest™ Rapid Readout Biological Indicator  
is removed from the 3M™ Attest™ Auto‑reader 390G well before a final result has been  
detected. Data will be lost if the 3M™ Attest™ Rapid Readout Biological Indicator is not  
replaced in the incubation well within 10 seconds.  
CARE AND CLEANING  
CAUTION: Always unplug the 3M™ Attest™ Auto‑reader 390G and allow to  
cool before cleaning.  
Do not spill liquid onto or into the instrument. Do not immerse the unit in liquid.  
Clean external surfaces using only the instructions provided by the manufacturer.  
The instrument must be returned to the manufacturer for repair.  
Alarm On/Off  
The audible alarm for positive results can be turned off for a particular 3M™ Attest™  
Rapid Readout Biological Indicator. This is accomplished by pressing the “Buzzer Alarm  
OFF” button after the alarm has sounded. This will turn off the alarm for that BI only. Any  
positive 3M™ Attest™ Rapid Readout Biological Indicator results subsequently detected  
in other wells will result in the alarm sounding. Removing the 3M™ Attest™ Rapid  
Readout Biological Indicator will reactivate the alarm for that well. Pressing the “Buzzer  
Alarm OFF” button will not silence the alarm for error or caution code alarm warnings. To  
disable the alarm, see configuration section.  
Clean the exterior surface of the 3M™ Attest™ Auto‑reader 390G by wiping it with a  
damp cloth moistened with a mild detergent. Do not attempt to clean interior components.  
If additional cleaning of the wells is required, call for repair or replacement.  
Note: The 3M™ Attest™ Auto‑reader 390G does not contain any user serviceable parts  
or require any routine maintenance. Any malfunctions will be indicated by an error code.  
Refer to the Troubleshooting Guide section for error code explanations.  
Stackable  
To save counter space, 3M™ Attest™ Auto‑reader 390Gs may be stacked two high.  
Information Button  
When the Information button is pressed once the 3M™ Attest™ Auto‑reader 390G  
displays two lines of text in the LCD display.  
The top row indicates (left to right):  
3M™ Attest™ Auto‑reader 390G serial number  
3M™ Attest™ Auto‑reader 390G embedded software version (V X.Y.Z)  
Current date and time.  
The bottom row indicates (left to right):  
The BI type configuration for wells 1‑5.  
The BI type configuration for wells 6‑10.  
Press the information button twice and the 3M™ Attest™ Auto‑reader 390G LCD panel  
will display two additional lines of text.  
The top row indicates:  
Dynamic IP address (if unit is connected to an Ethernet jack).  
The bottom row indicates (left to right):  
Date (YYYY‑MM‑DD) and time that power was supplied to the 3M™ Attest™  
Auto‑reader 390G.  
Temperature Monitoring  
The 3M™ Attest™ Auto‑reader 390G contains an internal temperature self‑diagnostic  
program. Should the temperature fall outside the specified range (37ºC +/‑2ºC) an error  
code will be displayed. The heater block temperature may be checked by depressing the  
Temperature button once.  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
TROUBLESHOOTING GUIDE  
In the event of an error code, unplug the 3M™ Attest™ Auto‑reader 390G and plug it back in to initiate the self‑diagnostic test. If error code re‑occurs, call for repair or replacement.  
The LCD display panel will indicate detected malfunctions by displaying a caution or error code under affected BI wells. The audible alarm, if active, sounds and the LCD display  
backlight flashes to alert the user to all caution (except C1) and error codes. The alarm and backlight can be silenced and extinguished, respectively, by pressing the Buzzer Alarm OFF  
button  
. Note that disabling the audible alarm does not disable the flashing LCD display backlight. Refer to the following:  
Code/Issue  
Problem  
Solution  
Unit Warm‑up Caution  
Allow a 30 minute warm‑up after 3M™ Attest™ Auto‑reader 390G is plugged in. (Note: 3M™ Attest™  
Rapid Readout BI readings are disabled until incubator block reaches 37ºC +/‑ 2ºC).  
C1  
Incubation block not at temperature.  
BI Removed from Well Caution  
Replace 3M™ Attest™ Rapid Readout BI in proper well within 10 seconds to prevent loss of collected  
data.  
C2  
C4  
E1  
3M™ Attest™ Rapid Readout BI has been removed before  
incubation is complete.  
Ambient Light Caution  
Move 3M™ Attest™ Auto‑reader 390G to reduced ambient light conditions.  
Temperature Control Error  
Unplug the unit and plug it back in to clear error.  
If error doesn’t clear, call for repair or replacement.  
This error will occur if system is no longer able to control  
the temperature of the incubator block (37ºC +/‑ 2ºC).  
Memory Error  
Unplug the unit and plug it back in to clear error.  
If error doesn’t clear, call for repair or replacement.  
E3  
E7  
E8  
3M™ Attest™ Auto‑reader 390G unable to save/read data  
from memory.  
Communications Error  
Call for repair or replacement.  
This error will occur if internal communications have failed.  
LED Error  
Unplug the unit and plug it back in to clear error.  
This error occurs for a specific well if the system detects  
that a UV LED or Optics Detection circuit is not working.  
If error doesn’t clear, call for repair or replacement. Do not use affected incubation well.  
BI Inserted Incorrectly Error  
Ensure BI is designed to be incubated and read in the 3M™ Attest™ Auto‑reader 390G.  
Ensure BI is fully inserted into incubation well.  
E11  
E12  
Incubation time remaining will not display and fluorescent  
readings are not taken.  
Power Failure Error  
Ensure power supply connection to Auto‑reader is secure.  
Power to Auto‑reader was off for too long to provide valid  
result for affected BI(s).  
Ambient Light Error (occupied well)  
Fluorescent result will not display.  
Unwanted alarm sounds when positive BI detected.  
Alarm Sounds  
Rely on optional visual pH color change result for BI in affected well. Move 3M™ Attest™  
Auto‑reader 390G to reduced ambient light conditions to avoid future occurrences.  
E14  
Other  
Disable alarm; refer to 3M™ Attest™ Auto‑reader 390G Configuration section for instructions.  
Replace 3M™ Attest™ Rapid Readout Biological Indicator in incubation/reader well within 10 seconds  
to prevent loss of data collected.  
Other  
3M™ Attest™ Rapid Readout Biological Indicator has been  
removed from incubation/reader well prematurely.  
Other  
Other  
LCD panel does not display properly.  
Negative Control Vial  
Unplug the unit and plug it back in. If problem persists call for repair or replacement.  
Ensure BI cap is closed, and glass ampoule is crushed. Tap BI until media flows to and wets spore strip  
at bottom of BI.  
BI not properly activated/ media did not contact spore strip.  
Negative Control Vial  
Other  
Other  
Check process indicator on BI label. Use a non‑sterilized vial as a control.  
Do not place tape or labels on vial.  
Control vial was sterilized.  
Unexplained Positive BI  
Adhesive residue on vial.  
Unexplained Positive BI  
Other  
Place the vial so it does not come in direct contact with chemical indicators or tape.  
Test vial is absorbing fluorescent residue from a chemical  
indicator or tape.  
3M™ Attest™ Auto‑reader 390G will not enter  
Configuration mode.  
On 3M™ Attest™ Auto‑reader 390G Web App Configuration Page, go to Pushbutton Configuration and  
Other  
Other  
select the Enable radio button. Save changes to begin using new setting.  
Pushbutton Configuration may be disabled on  
3M™ Attest™ Auto‑reader 390G Web App.  
3M™ Attest™ Auto‑reader 390G LCD display flashes  
and asterisks (***) and/or E11 and/or C4 codes appear on  
display below unoccupied incubation wells.  
Move 3M™ Attest™ Auto‑reader 390G away from sunlight or sources of incandescent light.  
8
Download from Www.Somanuals.com. All Manuals Search And Download.  
OPTIONAL 3M™ ATTEST™ AUTO‑READER 390G WEB APP  
Web Browser Requirements  
About the 3M™ Attest™ Auto‑reader 390G Web App Pages  
The 3M™ Attest™ Auto‑reader 390G Web App includes four pages:  
a) Status page  
The Web Interface is designed to be viewed in Internet Explorer 8.0 or higher. Earlier  
versions of Internet Explorer or other browsers may result in some web pages being  
displayed in a less‑than‑optimum manner. For best results, switching to Internet Explorer  
8.0 or higher is recommended. For full functionality, the use of an HTML5 compliant  
browser with JavaScript enabled is recommended.  
3M™ Attest™ Auto-reader 390G  
Connecting to the 3M™ Attest™ Auto‑reader 390G Web App  
Status  
History  
To use the 3M™ Attest™ Auto‑reader 390G Web App, the 3M™ Attest™ 390G  
Auto‑reader must be connected to the facility network using the supplied Ethernet  
cable. Consult with your facility IT personnel for assistance with network connectivity.  
Users may connect to the 3M™ Attest™ Auto‑reader 390G Web App using a URL based  
upon either the 3M™ Attest™ Auto‑reader 390G serial number or the IP Address. To  
view the 3M™ Attest™ Auto‑reader 390G Web App from a wireless device such as a  
mobile hand‑held device or a computer on a sub‑net different from that to which the  
3M™ Attest™ Auto‑reader 390G is connected, use the IP Address URL.  
ATTEST 1294  
ATTEST 1294  
1
2
3
4
5
6
7
8
9
10  
+
85  
240  
2012-10-31 12:32  
100122  
SN  
1. User‑entered nickname (optional)  
2. Well Configuration (1‑5 and 6‑10)  
3. Well Number  
4. BI Result/Remaining Incubation Time  
5. Date and Time  
Note: To access the 3M™ Attest™ Auto‑reader 390G Web App within a secure network,  
the PC or mobile device must have access to the secure network; options include  
connecting via a network Ethernet cable, signing in to the facility’s wireless network, or  
connecting via a virtual private network (VPN).  
A. Serial number URL  
6. 3M™ Attest™ Auto‑reader 390G Serial Number  
1. Connect the supplied Ethernet cable from the 3M™ Attest™ Auto‑reader 390G to an  
Ethernet jack.  
The Status page provides a real‑time representation of the 3M™ Attest™  
Auto‑reader 390G LCD panel display. The current date and time and 3M™ Attest™  
Auto‑reader 390G serial number are also indicated on this page. From this page, the user  
may navigate to:  
2. Depress the  
button once to display the 3M™ Attest™ Auto‑reader 390G 6‑digit  
serial number on the left side of the top row of text on the LCD display.  
the History page by clicking on History on the left of the screen.  
the Results Detail page by clicking on the well number of an occupied well (i.e., a well  
that currently has an incubating BI). Clicking on an unoccupied well takes the user to  
the History page.  
attest401008) from Step 2 in PC web browser address bar and press Go.  
4. 3M™ Attest™ Auto‑reader 390G Web App Status page will display on PC. For future  
ease of use, add this page to your web browser “Favorites.”  
b) Results Detail page  
B. IP Address URL  
The 3M™ Attest™ Auto‑reader 390G Web App may also be accessed using either a  
dynamic or static IP address.  
3M™ Attest™ Auto-reader 390G Results Detail  
Status  
History  
Print  
1. Dynamic IP Address  
A dynamic IP address is assigned when the Ethernet cable is connected from the  
3M™ Attest™ Auto‑reader 390G to a live Ethernet jack.  
Result  
Start Time  
2012-10-29 11:08:04  
#3 1294  
100122  
Well# and Configuration  
390G Serial Number  
Control?  
BI Lot#  
Press and release the  
button twice to display the 3M™ Attest™ Auto‑reader 390G’s  
Load#  
address bar of the PC or mobile hand‑held device and press Enter. The 3M™ Attest™  
Auto‑reader 390G Web App Status page will display.  
Sterilizer#  
Technician  
This page provides an individual record for an incubated BI. For complete documentation,  
the user may print a hard copy and/or save the record as an electronic file.  
2. Static IP address  
Some facility IT personnel may require devices connecting to the network to use a Static  
IP address. If this is the case at your facility, contact the IT department to request a Static  
IP address.  
Note 1: The top four rows (highlighted in green) are populated automatically by the  
3M™ Attest™ Auto‑reader 390G and cannot be changed by the user. The remaining rows  
are completed by the user to satisfy the documentation requirements of their facility.  
Note 2: The Results Detail page refreshes less frequently that the Status page. Until the BI  
result is available, the Results Detail page displays remaining minutes of incubation in the  
results field.  
Record the static IP address provided by your facility IT department here: ___________.  
You may wish to label the Ethernet jack with the assigned address. Enter  
PC or mobile hand‑held device and press Enter. The 3M™ Attest™ Auto‑reader 390G  
Web App Status page will display.  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
c) History page  
Important! Data is only stored for the last 100 incubated BIs.  
Embedded software in the 3M™ Attest™ Auto‑reader 390G saves the data from the  
auto‑populated (i.e., green highlighted) fields. The application relies on the temporary  
storage in the user’s web browser to save the data in the user‑populated fields. In both  
cases, data is only stored for the last 100 records, i.e., the 101st incubated BI will erase  
the information saved for the 1st incubated BI. To retain the Results Detail information for  
future reference it is best practice to print the page or save to an electronic file as soon as  
the BI result is available.  
3M™ Attest™ Auto-reader 390G  
Result  
Status  
History  
ID Well# and Configuration  
Start Time  
1
2
3
4
5
6
7
8
#4 1294  
#3 1294  
#3 1294  
#4 1294  
#5 1294  
#1 1294  
#4 1294  
#4 1294  
2012-10-30 12:31:25 ---  
2012-10-30 09:57:01 ---  
2012-10-30 07:51:51  
2012-10-30 16:15:55  
2012-10-30 16:01:07  
2012-10-30 15:58:18  
2012-10-30 11:52:27  
+
+
Using the BI History Page  
2012-10-30 11:42:26 C2  
The 3M™ Attest™ Auto‑reader 390G stores results for the last 100 incubated BIs on  
embedded software. The 3M™ Attest™ Auto‑reader 390G Web App History page provides  
a summary of these last 100 results.  
3M™ Attest™ Auto-reader 390G  
Status  
History  
ID Well# and Configuration  
Start Time  
Result  
1
2
3
4
5
6
7
8
#4 1294  
#3 1294  
#3 1294  
#4 1294  
#5 1294  
#1 1294  
#4 1294  
#4 1294  
2012-10-30 12:31:25 ---  
2012-10-30 09:57:01 ---  
d) Configuration page  
This page allows the configuration of the 3M™ Attest™ Auto‑reader 390G’s network and  
language settings from the 3M™ Attest™ Auto‑reader 390G Web App.  
2012-10-30 07:51:51  
2012-10-30 16:15:55  
2012-10-30 16:01:07  
2012-10-30 15:58:18  
2012-10-30 11:52:27  
+
+
Documenting BI Results using the 3M™ Attest™ Auto‑reader 390G  
Web App  
2012-10-30 11:42:26 C2  
To view or print an individual record, click on the ID number of the desired record to  
navigate to the Results Detail page for that BI.  
Note 1: Column 1, “ID” is the assigned ID number with ID 1 representing the newest  
record and ID 100 representing the oldest record. For convenience, all columns are  
sortable; to help find a particular record, simply click once on the column heading to sort  
in ascending order and click a second time to reverse the order.  
3M™ Attest™ Auto-reader 390G  
Status  
History  
ATTEST 1294  
ATTEST 1294  
1
2
3
4
5
6
7
8
9
10  
Note 2: The History page refreshes less frequently than the Status page. Until the BI result  
is available, the History page displays ••• in the Results column indicating incubation is  
still in‑process.  
229  
231  
2012-10-30 16:06  
100122  
SN  
1. Open the Results Detail page  
After inserting a new 3M™ Attest™ Rapid Readout BI in the 3M™ Attest™  
Auto‑reader 390G, it is a best practice to open the 3M™ Attest™ Auto‑reader 390G  
Web App Results Detail page for that BI by clicking on the appropriate well number  
from the Status page.  
2. Enter the specific information for that BI (e.g., BI Lot #, sterilizer/load number,  
Technician name, etc.) in the provided editable fields. Use the Notes section to capture  
information not prescribed in the provided row titles; for example, you may wish to use  
this field to record load items.  
3M™ Attest™ Auto-reader 390G Results Detail  
Status  
History  
Print  
Result  
223  
Start Time  
2012-10-30 15:58:18  
Well# and Configuration  
390G Serial Number  
#1 1294  
100122  
Control?  
BI Lot#  
No  
2014-11AD  
Load#  
1
Sterilizer#  
Technician  
2
sdf  
3. The entered information is saved automatically to temporary web browser storage on  
the user’s PC every 60 seconds, or by leaving the Results Details page.  
4. Print/File. Once the BI result is available, print or file the Results Detail page to create  
a record. The page may be physically printed and/or saved as an electronic file.  
10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Using the Configuration Page  
Certain set‑up and configuration options are available on the password‑protected 3M™ Attest™ Auto‑reader 390G Web App Configuration page. Ideally, the 3M™ Attest™  
Auto‑reader 390G should be empty (i.e., have no incubating BIs) when it is reconfigured. Navigate to the Configuration page by entering either:  
in the web browser address bar and pressing Go or Enter. A security screen will appear. Fill in the fields as follows:  
User name: manager  
Password: Attest  
and press OK.  
Click on the desired radio buttons to change settings.  
Selected setting indicates language in which text on 3M™ Attest™ Auto‑reader 390G Web App pages will appear on the PC/mobile device  
screen.  
Language  
Selecting Enable allows the user to configure the 3M™ Attest™ Auto‑reader 390G from the buttons on the unit itself.  
Pushbutton Configuration  
Selecting Disable prevents the user from being able to configure the 3M™ Attest™ Auto‑reader 390G from the buttons on the unit  
itself.  
Selecting Enable enables the audible alarm. The speaker icon will appear on the 3M™ Attest™ Auto‑reader 390G LCD panel.  
Buzzer Configuration  
Selecting Disable disables the audible alarm. The speaker icon will disappear from the 3M™ Attest™ Auto‑reader 390G LCD panel.  
Naming the 3M™ Attest™ Auto‑reader 390G may be useful in facilities with multiple units. The assigned nickname will appear on the  
3M™ Attest™ Auto‑reader 390G Web App Status page.  
Attest Auto‑reader 390G Nickname  
Default: Setting is checked indicating the Web App will use the Dynamic IP address automatically assigned when the 3M™ Attest™  
Enable DHCP  
Auto‑reader 390G is connected to the network. Remove the check if you wish to use a Static IP address assigned by your facility IT.  
IP Address  
Enter Static IP address assigned by IT.  
Gateway  
Enter Gateway.  
Subnet Mask  
Primary DNS  
Use Computer Time  
Enter Subnet Mask.  
Enter Primary DNS.  
Check this box if you wish to set the clock on‑board the 3M™ Attest™ Auto‑reader 390G using the computer’s clock.  
Alternatively, leave the Use Computer Time box unchecked and complete the date and time boxes to set the  
3M™ Attest™ Auto‑reader 390G clock.  
Attest Auto‑reader 390G Time  
After changing desired settings, click on Save Config button to save your changes. A new screen confirming the settings were successfully changed appears.  
11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
TROUBLESHOOTING THE 3M™ ATTEST™ AUTO‑READER 390G WEB APP  
Issue  
Problem  
Solution  
Ensure 3M™ Attest™ Auto‑reader 390G Ethernet cable is connected to a live jack and that  
connection to 3M™ Attest™ Auto‑reader 390G is secure.  
Unable to access 3M™ Attest™  
Auto‑reader 390G Web App on  
networked computer  
Ensure computer is connected to network.  
3M™ Attest™ Auto‑reader 390G and/or  
computer not connected to network.  
If failure is with connection attempted using 3M™ Attest™ Auto‑reader 390G serial number  
URL, try connecting using IP address URL.  
Contact facility IT or 3M if problem persists.  
3M™ Attest™ Auto‑reader 390G not  
connected to network.  
Ensure 3M™ Attest™ Auto‑reader 390G Ethernet cable is connected to a live jack and that  
connection to 3M™ Attest™ Auto‑reader 390G is secure.  
Unable to access 3M™ Attest™  
Auto‑reader 390G Web App on  
mobile device  
For security reasons, facility firewall may  
prevent mobile devices not connected to  
the facility network from connecting to  
secure IP addresses.  
Contact facility IT for their recommendation on connecting to the secure network. Solutions may  
include enabling Wi‑Fi on the mobile device and signing into the facility network or connecting  
to the facility virtual private network (VPN) on the mobile device.  
User entered data is written to temporary  
web browser storage. Browser may have  
cleared data.  
Set web browser to accept (allow or enable) cookies.  
Data entered by user on Results  
Detail page disappears  
As a best practice, print or save Results Detail record as soon as BI result is final.  
BI result displayed on 3M™ Attest™  
Auto‑reader 390G LCD panel does  
not appear on Results Detail page  
The Results Detail page has not  
refreshed yet.  
Use your browser to refresh the Results Detail page, or wait up to 60 seconds for the page to  
automatically refresh.  
The Results Detail page only opens  
from the Status page for occupied  
3M™ Attest™ Auto‑reader 390G wells.  
Results Detail page will not open  
from the Status page  
Do not attempt to navigate to the Results Detail page from unoccupied wells on the Status page.  
REPAIR AND REPLACEMENT  
In the U.S.A., 3M™ Attest™ Auto‑reader 390Gs must be repaired at:  
3M Health Care Service Center  
Suite 200, Bldg. 502  
3350 Granada Avenue North  
Oakdale, MN 55128  
1‑800‑292‑6298  
Fax: 1‑800‑770‑8016  
Outside of the U.S.A., contact your local 3M Subsidiary.  
If you have questions, call our Customer Service number: 1‑800‑292‑6298.  
Made in U.S.A. by  
3M Health Care  
2510 Conway Ave  
St. Paul, MN 55144  
1‑800‑228‑3957  
© 2013, 3M. All rights reserved.  
3M and Attest are trademarks of 3M.  
Used under license in Canada.  
12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Made in U.S.A. by  
3M Health Care  
© 2013, 3M. All rights reserved.  
3M and Attest are trademarks of 3M.  
Used under license in Canada.  
EC REP  
3M Deutschland GmbH  
Health Care Business  
Carl-Schurz-Str. 1  
41453 Neuss  
2510 Conway Ave.  
St. Paul, MN 55144  
1-800-228-3957  
Issue Date: 2013-08  
34-8712-0286-6  
Germany  
Download from Www.Somanuals.com. All Manuals Search And Download.  

3Com Modem 3CP3617B User Manual
3Com Network Router 355094 User Manual
3D Innovations Refrigerator NC60 User Manual
Acnodes Personal Computer FPC 8057 User Manual
Agilent Technologies Yard Vacuum 86140B User Manual
Airflow Systems Carbon Monoxide Alarm IAQ50 User Manual
Alliance Laundry Systems Washer SF18VNV User Manual
American Audio CD Player DCD PRO500 User Manual
AOSafety Respiratory Product Full Face Respirator Mask User Manual
Tripp Lite Surge Protector Protect It User Manual