Your new Measurement Computing product comes with a fantastic extra —
Management committed to your satisfaction!
Refer to www.mccdaq.com/execteam.html for the names, titles, and contact information of each key executive at Measurement
Computing.
Thank you for choosing a Measurement Computing product—and congratulations! You own the finest, and you can now enjoy
the protection of the most comprehensive warranties and unmatched phone tech support. It’s the embodiment of our mission:
To provide PC-based data acquisition hardware and software that will save time and save money.
Simple installations minimize the time between setting up your system and actually making measurements. We offer quick and
simple access to outstanding live FREE technical support to help integrate MCC products into a DAQ system.
Lifetime warranty: Every hardware product manufactured by Measurement Computing Corporation is warranted against
defects in materials or workmanship for the life of the product. Products found defective are repaired or replaced promptly.
Lifetime Harsh Environment Warranty®: We will replace any product manufactured by Measurement Computing
Corporation that is damaged (even due to misuse) for only 50% of the current list price. I/O boards face some tough operating
conditions, some more severe than the boards are designed to withstand. When a board becomes damaged, just return the unit
with an order for its replacement at only 50% of the current list price. We don’t need to profit from your misfortune. By the way,
we honor this warranty for any manufacturer’s board that we have a replacement for.
30 Day Money Back Guarantee: You may return any Measurement Computing Corporation product within 30 days of
purchase for a full refund of the price paid for the product being returned. If you are not satisfied, or chose the wrong product by
mistake, you do not have to keep it. Please call for an RMA number first. No credits or returns accepted without a copy of the
original invoice. Some software products are subject to a repackaging fee.
These warranties are in lieu of all other warranties, expressed or implied, including any implied warranty of merchantability or
fitness for a particular application. The remedies provided herein are the buyer’s sole and exclusive remedies. Neither
Measurement Computing Corporation, nor its employees shall be liable for any direct or indirect, special, incidental or
consequential damage arising from the use of its products, even if Measurement Computing Corporation has been notified in
advance of the possibility of such damages.
HM USB-1616HS-2.doc
3
Trademark and Copyright Information
TracerDAQ, Universal Library, Harsh Environment Warranty, Measurement Computing Corporation, and the Measurement
Computing logo are either trademarks or registered trademarks of Measurement Computing Corporation.
Windows, Microsoft, and Visual Studio are either trademarks or registered trademarks of Microsoft Corporation
LabVIEW is a trademark of National Instruments.
CompactFlash is a registered trademark of SanDisk Corporation.
XBee and XBee-PRO are trademarks of MaxStream, Inc.
All other trademarks are the property of their respective owners.
Information furnished by Measurement Computing Corporation is believed to be accurate and reliable. However, no
responsibility is assumed by Measurement Computing Corporation neither for its use; nor for any infringements of patents or
other rights of third parties, which may result from its use. No license is granted by implication or otherwise under any patent or
copyrights of Measurement Computing Corporation.
All rights reserved. No part of this publication may be reproduced, stored in a retrieval system, or transmitted, in any form by any
means, electronic, mechanical, by photocopying, recording, or otherwise without the prior written permission of Measurement
Computing Corporation.
Notice
Measurement Computing Corporation does not authorize any Measurement Computing Corporation product for use
in life support systems and/or devices without prior written consent from Measurement Computing Corporation.
Life support devices/systems are devices or systems which, a) are intended for surgical implantation into the body,
or b) support or sustain life and whose failure to perform can be reasonably expected to result in injury.
Measurement Computing Corporation products are not designed with the components required, and are not subject
to the testing required to ensure a level of reliability suitable for the treatment and diagnosis of people.
4
Table of Contents
Preface
Conventions used in this user's guide.................................................................................................................7
Where to find more information.........................................................................................................................7
Chapter 1
Introducing the USB-1616HS-2............................................................................................................8
Software features................................................................................................................................................8
Chapter 2
Installing the USB-1616HS-2................................................................................................................9
Configuring the hardware.................................................................................................................................12
Chapter 3
Analog input.....................................................................................................................................................19
Tips for making accurate temperature measurements......................................................................................................22
Analog output...................................................................................................................................................23
Example: Analog channel scanning of voltage inputs and streaming analog outputs......................................................23
Digital input scanning......................................................................................................................................................24
Triggering.........................................................................................................................................................24
Software-based triggering................................................................................................................................................25
Stop trigger modes...........................................................................................................................................................26
5
USB-1616HS-2 User's Guide
Tips for making high-speed counter measurements (> 1 MHz).......................................................................................27
Counter modes.................................................................................................................................................................27
Detection setpoint details.................................................................................................................................................42
Chapter 4
Calibrating the USB-1616HS-2...........................................................................................................44
Chapter 5
Specifications......................................................................................................................................45
Analog input.....................................................................................................................................................45
Accuracy..........................................................................................................................................................................45
Triggering.........................................................................................................................................................49
Declaration of Conformity..................................................................................................................53
6
Preface
About this User's Guide
What you will learn from this user's guide
This user's guide explains how to install, configure, and use the USB-1616HS-4 so that you get the most out of
its analog I/O, thermocouple (TC) input, digital I/O, counter/timer I/O features.
This user's guide also refers you to related documents available on our web site, and to technical support
resources.
Conventions used in this user's guide
For more information on …
Text presented in a box signifies additional information and helpful hints related to the subject matter you are
reading.
Caution! Shaded caution statements present information to help you avoid injuring yourself and others,
damaging your hardware, or losing your data.
<#:#>
Angle brackets that enclose numbers separated by a colon signify a range of numbers, such as those assigned
to registers, bit settings, etc.
bold text
Bold text is used for the names of objects on the screen, such as buttons, text boxes, and check boxes. For
example:
1. Insert the disk or CD and click the OK button.
italic text
Italic text is used for the names of manuals and help topic titles, and to emphasize a word or phrase. For
example:
Never touch the exposed pins or circuit connections on the board.
Where to find more information
The following electronic documents provide information that can help you get the most out of your USB-
1616HS-2.
MCC's Specifications: USB-1616HS-2 (the PDF version of the "Specifications" chapter in this guide) is
MCC's Quick Start Guide is available on our web site at
MCC's Guide to Signal Connections is available on our web site at
MCC's Universal Library User's Guide is available on our web site at
MCC's Universal Library Function Reference is available on our web site at
MCC's Universal Library for LabVIEW™ User’s Guide is available on our web site at
USB-1616HS-2 User's Guide (this document) is also available on our web site at
7
Chapter 1
Introducing the USB-1616HS-2
Overview: USB-1616HS-2 features
The USB-1616HS-2 is supported under popular Microsoft® Windows® operating systems. The USB-1616HS-2
board is a multifunction measurement and control board designed for the USB bus.
The USB-1616HS-2 provides either eight differential or 16 single-ended analog inputs with 16-bit resolution. It
offers seven software-selectable analog input ranges of ±10 V, ±5 V, ±2 V, ±1 V, ±0.5 V, ±0.2 V, and ±0.1V.
You can configure up to eight of the analog inputs as differential thermocouple (TC) inputs.
The USB-1616HS-2 has two 16-bit, 1 MHz analog output channels with an output range of -10 V to +10 V.
The board has 24 high-speed lines of digital I/O, two timer outputs, and four 32-bit counters. It provides up to
4 MHz scanning on all digital input lines1.
Six banks of removable screw-terminal blocks provide connectivity to the analog input channels, digital I/O
lines, counter/timer channels, and analog outputs.
You can operate all analog I/O, digital I/O, and counter/timer I/O synchronously.
Software features
For information on the features of InstaCal and the other software included with your USB-1616HS-2, refer to
the Quick Start Guide that shipped with your device. The Quick Start Guide is also available in PDF at
1 Higher rates—up to 12 MHz—are possible depending on the platform and the amount of data being transferred.
8
Chapter 2
Installing the USB-1616HS-2
What comes with your USB-1616HS-2 shipment?
As you unpack your USB-1616HS-2, verify that the following components are included.
Hardware
USB-1616HS-2
USB cable (2-meter length)
TR-2U power supply and CA-1* line cord
AC-to-DC conversion power supply and cord plugs into the external power connector of the USB-1616HS-2.
* European customers: Contact Measurement Computing to order the CA-261 line cord for your region.
Optional components
Expansion devices and cables and that are compatible with the USB-1616HS-2 and must be ordered separately.
If you ordered any of the following products with your device, they should be included with your shipment.
9
USB-1616HS-2 User's Guide
Installing the USB-1616HS-2
AI-EXP48
Analog input expansion module adds up to 24 differential or 48 single-ended inputs to the USB-1616HS-2.
CA-96A expansion cable
Expansion cable for connecting to the AI-EXP48 expansion board.
Additional documentation
In addition to this hardware user's guide, you should also receive the Quick Start Guide (available in PDF at
www.mccdaq.com/PDFmanuals/DAQ-Software-Quick-Start.pdf). This booklet supplies a brief description of
the software you received with your USB-1616HS-2 and information regarding installation of that software.
Please read this booklet completely before installing any software or hardware.
Unpacking the USB-1616HS-2
As with any electronic device, you should take care while handling to avoid damage from static
electricity. Before removing the USB-1616HS-2 from its packaging, ground yourself using a wrist strap or by
simply touching the computer chassis or other grounded object to eliminate any stored static charge.
If any components are missing or damaged, notify Measurement Computing Corporation immediately by
phone, fax, or e-mail:
Phone: 508-946-5100 and follow the instructions for reaching Tech Support.
Fax: 508-946-9500 to the attention of Tech Support
Installing the software
Refer to the Quick Start Guide for instructions on installing the software on the Measurement Computing Data
Acquisition Software CD. This booklet is available in PDF at www.mccdaq.com/PDFmanuals/DAQ-Software-
We recommend that you download the latest Windows Update onto your computer before installing and
operating the USB-1616HS-2.
10
USB-1616HS-2 User's Guide
Installing the USB-1616HS-2
Installing the hardware
To connect the USB-1616HS-2 to your system, turn your computer on, and then do the following:
Connect signal lines to the USB-1616HS-2's removable screw terminal blocks.
o
o
Connect thermocouple signals as differential connections (see Figure 1). The negative (typically, the
red) thermocouple wire connects to the channel's LO connector, and the other color wire connects to
the channel's HI connector.
Always use differential input mode for thermocouple connections.
Figure 1. Single-ended voltage connections (V1 and V2) and differential thermocouple connections (V3)
If you are using an AI-EXP48 expansion device, connect it to the USB-1616HS-2. Make sure you do not
connect the AI-EXP48 to a live USB-1616HS-2. If the USB cable is connected to the computer, unplug it
before you connect the AI-EXP48.
If you are using the TR-2U external supply (sold separately), connect the power supply to the USB-
1616HS-2's external power connector, and plug the other end into a power outlet.
The TR-2U is optional, but can be used in any scenario. You may need a TR-2U power supply if the USB
port does not provide enough power for your USB-1616HS-2 application.
The USB-1616HS-2 requires 3000 mW by itself, and 3400 mW when connected to the AI-EXP48.
By USB2 standards, USB 2.0 ports are required to provide at least 2500 mW.
Connect the USB cable to the USB-1616HS-2 USB connector and to a USB port on your computer. A
USB2.0 port is recommended—connecting to a USB1.1 port results in lower performance.
When you connect the USB-1616HS-2 for the first time, a Found New Hardware message opens as the
USB-1616HS-2 is detected. When the message closes, the installation is complete.
The power LED (bottom LED) blinks during device detection and initialization, and then remains solid if
properly detected. If not, check if the USB-1616HS-2 has sufficient power. When the board is first powered on,
there is usually a momentary delay before the power LED begins to blink, or come on solid.
Caution! Do not disconnect any device from the USB bus while the computer is communicating with the
USB-1616HS-2, or you may lose data and/or your ability to communicate with the USB-1616HS-
2.
11
USB-1616HS-2 User's Guide
Installing the USB-1616HS-2
Configuring the hardware
All hardware configuration options on the USB-1616HS-2 are software-controlled. You can select some of the
configuration options using InstaCal, such as the analog input configuration (16 single-ended or 8 differential
channels), and the edge used for pacing when using an external clock. When measuring from thermocouples,
make sure you configure the channels for differential mode.
Once selected, any program that uses the Universal Library initializes the hardware according to these
selections.
Caution! Turn off power to all devices connected to the system before making connections. Electrical shock
or damage to equipment can result even under low-voltage conditions.
Information on signal connections
General information regarding signal connection and configuration is available in the Guide to Signal
Caution! Always handle components carefully, and never touch connector terminals or circuit components
unless you are following ESD guidelines in an appropriate ESD-controlled area. These guidelines
include using properly-grounded mats and wrist straps, ESD bags and cartons, and related
procedures.
Avoid touching board surfaces and onboard components. Only handle boards by their edges. Make
sure the USB-1616HS-2 does not come into contact with foreign elements such as oils, water, and
industrial particulate.
The discharge of static electricity can damage some electronic components. Semiconductor
devices are especially susceptible to ESD damage.
Connecting the board for I/O operations
Connectors, cables – main I/O connector
The following table lists the board connectors, applicable cables, and compatible accessory products for the
USB-1616HS-2.
Main connector specifications
Main connectors
Six banks of removable screw-terminal blocks
25-pin DSUB, female (DSUB25F)
CA-96A
Expansion connector
Compatible cable for the 25-pin expansion connector
Compatible accessory product for the 25-pin expansion
connector
AI-EXP48 expansion board with screw terminals (can connect
to the USB-1616HS-2 directly, or with the CA-96A cable)
12
USB-1616HS-2 User's Guide
Installing the USB-1616HS-2
Screw terminal pin outs
USB-1616HS-2 screw terminal pin out – single-ended connections
Analog common (Ad)
Analog output 0 (AO0)
Analog output 1 (AO1)
NC
Digital common (Dd)
FIRSTPORTA Bit 0 (A0)
FIRSTPORTA Bit 1 (A1)
FIRSTPORTA Bit 2 (A2)
FIRSTPORTA Bit 3 (A3)
FIRSTPORTA Bit 4 (A4)
FIRSTPORTA Bit 5 (A5)
FIRSTPORTA Bit 6 (A6)
FIRSTPORTA Bit 7 (A7)
Digital common (Dd)
Timer 0 (T0)
NC
Analog common (Ad)
CAL (Reserved for self-calibration)
Signal ground (Sd)
Analog Out
DIG-Tmr I/O
Digital common (Dd)
TTL trigger (TRG)
Output scan clock I/O (DPR)
Input scan clock I/O (APR)
Timer 1 (T1)
Analog common (Ad)
CH 0 (0H)
CH 8 (8L)
Analog common (Ad)
CH 1 (1H)
CH 9 (9L)
Analog common (Ad)
CH 2 (2H)
CH 10 (10L)
Analog common (Ad)
CH 3 (3H)
Digital common (Dd)
FIRSTPORTB Bit 0 (B0)
FIRSTPORTB Bit 1 (B1)
FIRSTPORTB Bit 2 (B2)
FIRSTPORTB Bit 3 (B3)
FIRSTPORTB Bit 4 (B4)
FIRSTPORTB Bit 5 (B5)
FIRSTPORTB Bit 6 (B6)
FIRSTPORTB Bit 7 (B7)
Digital common (Dd)
Analog In
Dig-Ctr I/O
Counter 0 (CT0)
Counter 1 (CT1)
CH 11 (11L)
Analog common (Ad)
CH 4 (4H)
CH 12 (12L)
Analog common (Ad)
CH 5 (5H)
CH 13 (13L)
Analog common (Ad)
CH 6 (6H)
CH 14 (14L)
Analog common (Ad)
CH 7 (7H)
Digital common (Dd)
FIRSTPORTC Bit 0 (C0)
FIRSTPORTC Bit 1 (C1)
FIRSTPORTC Bit 2 (C2)
FIRSTPORTC Bit 3 (C3)
FIRSTPORTC Bit 4 (C4)
FIRSTPORTC Bit 5 (C5)
FIRSTPORTC Bit 6 (C6)
FIRSTPORTC Bit 7 (C7)
Digital common (Dd)
Counter 2 (CT2)
Analog In
Dig-Ctr I/O
CH 15 (15L)
Counter 3 (CT3)
13
USB-1616HS-2 User's Guide
Installing the USB-1616HS-2
USB-1616HS-2 screw terminal pin out – differential connections
Analog common (Ab)
Analog output 0 (AO0)
Analog output 1 (AO1)
NC
Digital common (Db)
FIRSTPORTA Bit 0 (A0)
FIRSTPORTA Bit 1 (A1)
FIRSTPORTA Bit 2 (A2)
FIRSTPORTA Bit 3 (A3)
FIRSTPORTA Bit 4 (A4)
FIRSTPORTA Bit 5 (A5)
FIRSTPORTA Bit 6 (A6)
FIRSTPORTA Bit 7 (A7)
Digital common (Db)
Timer 0 (T0)
NC
Analog common (Ab)
CAL (Reserved for self-calibration)
Signal ground (Sb)
Digital common (Db)
TTL trigger (TRG)
Analog Out
DIG-Tmr I/O
Output scan clock I/O (DPR)
Input scan clock I/O (APR)
Timer 1 (T1)
Analog common (Ab)
CH 0 HI (0H)
CH 0 LO (8L)
Analog common (Ab)
CH 1 HI (1H)
CH 1 LO (9L)
Analog common (Ab)
CH 2 HI (2H)
CH 2 LO (10L)
Analog common (Ab)
CH 3 HI (3H)
Digital common (Db)
FIRSTPORTB Bit 0 (B0)
FIRSTPORTB Bit 1 (B1)
FIRSTPORTB Bit 2 (B2)
FIRSTPORTB Bit 3 (B3)
FIRSTPORTB Bit 4 (B4)
FIRSTPORTB Bit 5 (B5)
FIRSTPORTB Bit 6 (B6)
FIRSTPORTB Bit 7 (B7)
Digital common (Db)
Analog In
Dig-Ctr I/O
Counter 0 (CT0)
Counter 1 (CT1)
CH 3 LO (11L)
Analog common (Ab)
CH 4 HI (4H)
CH 4 LO (12L)
Analog common (Ab)
CH 5 HI (5H)
CH 5 LO (13L)
Analog common (Ab)
CH 6 HI (6H)
CH 6 LO (14L)
Analog common (Ab)
CH 7 HI (7H)
Digital common (Db)
FIRSTPORTC Bit 0 (C0)
FIRSTPORTC Bit 1 (C1)
FIRSTPORTC Bit 2 (C2)
FIRSTPORTC Bit 3 (C3)
FIRSTPORTC Bit 4 (C4)
FIRSTPORTC Bit 5 (C5)
FIRSTPORTC Bit 6 (C6)
FIRSTPORTC Bit 7 (C7)
Digital common (Db)
Counter 2 (CT2)
Analog In
Dig-Ctr I/O
CH 7 LO (15L)
Counter 3 (CT3)
DSUB25F expansion connector
13
1
25
14
Figure 2. DSUB25 expansion connector pin out
14
Chapter 3
Functional Details
This chapter contains detailed information on all of the features available from the board, including:
a diagram and explanations of physical board components
a functional block diagram
information on how to use the signals generated by the board
diagrams of signals using default or conventional board settings
USB-1616HS-2 components
These USB-1616HS-2 components are shown in Figure 4.
Six removable screw terminal blocks
One USB port
One external power connector
One 25-pin expansion connector
Two LED indicators ("Active" and "Power")
Analog output, calibration,
TTL trigger and pacer signal
Analog input, screw terminal
blocks
DSUB25 expansion
connector
Analog input, screw terminal
blocks
Figure 4. USB-1616HS-2 components – front view
16
USB-1616HS-2 User's Guide
Functional Details
Device "Active"
LED
Counter and digital I/O (port C)
screw terminal blocks
External
"Power" LED
Counter and digital I/O (port B)
screw terminal blocks
USB connector
Timer and digital I/O (port A)
screw terminal blocks
External power
connector
Figure 5. USB-1616HS-2 components – rear view
External power connector
Although the USB-1616HS-2 is powered by a USB port on a host PC, an external power connector may also be
required to provide sufficient power for the USB-1616HS-2.
Connect the optional TR-2U power supply to the external power supply connector. This power supply provides
9 VDC, 1 A power to the USB-1616HS-2.
17
USB-1616HS-2 User's Guide
Functional Details
USB-1616HS-2 block diagram
Figure 6 shows a simplified block diagram of the USB-1616HS-2. This board provides all of the functional
elements shown in the figure.
16-bit
1 MHz
D/A converters
output clock
2
8 diff/16 SE
analog inputs
Programmable gain
amplifier
Analog channel
input protection
MUX
A
x1, x2,
x5, x10, x20
x50, x100
16
2
16-bit, 1 MHz
A/D converter
One TTL
trigger input
A
One analog
input pacer clock
512-step
random access
channel/gain
sequencer
1 MHz
input clock
1 MSample
FIFO
Data Buffer
8
8
Two 16-bit
Timer outputs
Sequencer reset
2
Programmable
sequencer
timebase
Four 32-bit
counter inputs
2
2
1
µ
s to 6 hours
Three 8-bit
DIO ports
A
24
USB port
System
controller
USB
controller
DSUB25F
Expansion connector
External
power
Configurable
PLD
Configurable
EPROM
DC to DC
converter
Connect the optional
power supply
if the USB cannot
supply enough power.
Figure 6. USB-1616HS-2 functional block diagram
Synchronous I/O – mixing analog, digital, and counter scanning
The USB-1616HS-2 can read analog, digital, and counter inputs, while generating up to two analog outputs and
digital pattern outputs at the same time. Digital and counter inputs do not affect the overall A/D rate because
these inputs use no time slot in the scanning sequencer.
For example, one analog input channel can be scanned at the full 1 MHz A/D rate along with digital and counter
input channels. Each analog channel can have a different gain, and counter and digital channels do not need
additional scanning bandwidth as long as there is at least one analog channel in the scan group.
Digital input channel sampling is not done during the "dead time" of the scan period where no analog sampling
is being done either.
18
USB-1616HS-2 User's Guide
Functional Details
Analog input
The USB-1616HS-2 has a 16-bit, 1-MHz A/D coupled with 16 single-ended, or eight differential analog inputs.
Seven software programmable ranges provide inputs from ±10 V to ±100 mV full scale.
Analog input scanning
The USB-1616HS-2 has several scanning modes to address various applications. You can load the 512-location
scan buffer with any combination of analog input channels. All analog input channels in the scan buffer are
measured sequentially at 1 µs per channel by default.
For example, in the fastest mode, with ADC settling time set to 1 µs, a single analog channel can be scanned
continuously at 1 MS/s; two analog channels can be scanned at 500 kS/s each; 16 analog input channels can be
scanned at 62.5 kS/s.
Settling time
For most applications, leave the settling time at its default of 1 µs.
However, if you are scanning multiple channels, and one or more channels are connected to a high-impedance
source, you may get better results by increasing the settling time. Remember that increasing the settling time
reduces the maximum acquisition rate.
You can set the settling time to 1 µs, 5 µs, 10 µs, or 1 ms.
Example: Analog channel scanning of voltage inputs
Figure 7 shows a simple acquisition. The scan is programmed pre-acquisition and is made up of six analog
channels (Ch0, Ch1, Ch3, Ch4, Ch6, and Ch7). Each of these analog channels can have a different gain. The
acquisition is triggered and the samples stream to the PC. Each analog channel requires one microsecond of
scan time—therefore the scan period can be no shorter than 6 µs for this example. The scan period can be made
much longer than 6 µs—up to 1 s. The maximum scan frequency is one divided by 6 µs, or 166,666 Hz.
Figure 7. Analog channel scan of voltage inputs example
Example: Analog channel scanning of voltage and temperature inputs
Figure 8 shows a programmed pre-acquisition scan made up of six analog channels (Ch0, Ch1, Ch5, Ch11,
Ch12, Ch13). Each of these analog channels can have a different gain. You can program channels 0 and 1 to
directly measure TCs.
In this mode, oversampling is programmable up to 16384 oversamples per channel in the scan group. When
oversampling is applied, it is applied to all analog channels in the scan group, including temperature and voltage
channels. Digital channels are not oversampled.
19
USB-1616HS-2 User's Guide
Functional Details
If you want 256 oversamples, then each analog channel in the scan group will take 256 µs, and the returned 16-
bit value represents an average of 256 consecutive 1 µs samples of that channel. The acquisition is triggered and
16-bit values—each representing an average of 256—stream to the PC via the USB cable. Since two of the
channels in the scan group are temperature channels, you need the acquisition engine to read a cold-junction-
compensation (CJC) temperature every scan.
Figure 8. Analog channel scanning of voltage and temperature inputs example
Since the targeted number of oversamples is 256 in this example, each analog channel in the scan group
requires 256 microseconds to return one 16-bit value. The oversampling is also done for CJC temperature
measurement channels, making the minimum scan period for this example 7 x 256 µs, or 1792 µs. The
maximum scan frequency is the inverse of this number, 558 Hz.
For accurate measurements, you must associate TC and CJC channels properly
The TC channels must immediately follow their associated CJC channels in the channel array. For accurate TC
readings, associate CJC0 with TC0, CJC1 with TC1 and TC2, CJC2 with TC3, CJC3 with TC4, CJC4 with TC5
and TC6, and CJC5 with TC7.
When the AI-EXP48 module is connected to the USB-1616HS-2, associate CJC6 with TC8 through TC11,
CJC7 with TC12 through TC15, CJC8 with TC16 through TC19, CJC9 with TC20 through TC23, CJC10 with
TC24 through TC27, and CJC11 with TC28 through TC31.
Example: Analog and digital scanning, once per scan mode
The scan is programmed pre-acquisition and is made up of six analog channels (Ch0, Ch2, Ch5, Ch11, Ch13,
Ch15) and four digital channels (16-bits of digital IO, three counter inputs.) Each of the analog channels can
have a different gain.
The acquisition is triggered and the samples stream to the PC via the USB cable. Each analog channel requires
one microsecond of scan time. Therefore, the scan period can be no shorter than 6 µs for this example. All of
the digital channels are sampled at the start of scan and do not require additional scanning bandwidth as long as
there is at least one analog channel in the scan group. The scan period can be made much longer than 6 µs, up to
1 second. The maximum scan frequency is one divided by 6 µs or 166,666 Hz.
Figure 9. Analog and digital scanning, once per scan mode example
20
USB-1616HS-2 User's Guide
Functional Details
The counter channels may return only the lower 16-bits of count value if that is sufficient for the application.
They could also return the full 32-bit result if necessary. Similarly, the digital input channel could be the full 24
bits if desired or only eight bits if that is sufficient. If the three counter channels are all returning 32-bit values
and the digital input channel is returning a 16-bit value, then 13 samples are being returned to the PC every scan
period, with each sample being 16-bits. The 32-bit counter channels are divided into two 16-bit samples—one
for the low word, and the other for the high word. If the maximum scan frequency is 166,666 Hz, then the data
bandwidth streaming into the PC is 2.167 MS/s. Some slower PCs may have a problem with data bandwidths
greater than 6 MS/s.
The USB-1616HS-2 has an onboard 1 MS buffer for acquired data.
Example: Sampling digital inputs for every analog sample in a scan group
The scan is programmed pre-acquisition and is made up of six analog channels (Ch0, Ch2, Ch5, Ch11, Ch13,
Ch15) and four digital channels (16-bits of digital input, three counter inputs.) Each of the analog channels can
have a different gain.
The acquisition is triggered and the samples stream to the PC via the USB cable. Each analog channel requires
one microsecond of scan time therefore the scan period can be no shorter than 6 µs for this example. All of the
digital channels are sampled at the start of scan and do not require additional scanning bandwidth as long as
there is at least one analog channel in the scan group. The 16-bits of digital input are sampled for every analog
sample in the scan group. This allows up to 1 MHz digital input sampling while the 1 MHz analog sampling
bandwidth is aggregated across many analog input channels.
The scan period can be made much longer than 6 µs—up to 1 second. The maximum scan frequency is one
divided by 6 µs, or 166,666 Hz. Note that digital input channel sampling is not done during the "dead time" of
the scan period where no analog sampling is being done either.
Figure 10. Analog and digital scanning, once per scan mode example
If the three counter channels are all returning 32-bit values and the digital input channel is returning a 1-bit
value, then 18 samples are returned to the PC every scan period, with each sample being 16-bits. Each 32-bit
counter channel is divided into two 16-bit samples—one for the low word and the other for the high word. If the
maximum scan frequency is 166,666 Hz, then the data bandwidth streaming into the PC is 3 MS/s. Some slower
PCs may have a problem with data bandwidths greater than 6 MS/s.
The USB-1616HS-2 has an onboard 1 MS buffer for acquired data.
21
USB-1616HS-2 User's Guide
Functional Details
Thermocouple input
You can configure up to eight analog inputs in differential mode on the USB-1616HS-2 to accept a
thermocouple (TC) input. Built-in cold-junction sensors are provided for each of the screw-terminal connectors,
and any TC type can be attached to any of the eight thermocouple channels.
When measuring TCs, the USB-1616HS-2 can operate in an averaging mode, taking multiple readings on each
channel, applying digital filtering and cold-junction compensation, and then converting the readings to
temperature.
As a result, the USB-1616HS-2 measures channels with TCs attached at a rate from 50 Hz to 10 kHz,
depending on how much over-sampling is selected.
Additionally, a rejection frequency can be specified in which over sampling occurs during one cycle of either
50 Hz or 60 Hz, providing a high level of 50 Hz or 60 Hz rejection.
The USB-1616HS-2 does not have open thermocouple detection.
Tips for making accurate temperature measurements
Set the rejection frequency to equal the line frequency.
Warm up the USB-1616HS-2 for 60 minutes—including TC wires—so that it is thermally stabilized. This
warm-up time enables the CJC thermistors to more accurately measure the junction at the terminal block.
Make sure the surrounding environment is thermally stabilized and ideally around 20 °C to 30 °C. If the
device's ambient temperature is changing due to a local heating or cooling source, then the TC junction
temperature may be changing and the CJC thermistor will have a larger error.
Use small-diameter, instrument-grade TC wire. Small diameter TC wire has less effect on the TC junction
at the terminal block because less heat is transferred from the ambient environment to the junction.
Use shielded TC wire (see "Shielding" below) with the shield connected to analog common to reduce noise.
The USB-1616HS-2 has several analog commons on the screw terminals.
You can also minimize the effect of noise by averaging readings (see "Averaging" below), or combining
both shielding and averaging.
Refer to "Screw terminal pin outs" section starting on page 13 for the locations of these analog common
screw terminals.
Make sure the USB-1616HS-2 is mounted on a flat surface.
Be careful to avoid loading down the digital outputs too heavily (>1 mA). Heavy load down causes
significant heat generation inside the unit and increase the CJC thermistor error.
Shielding
Use shielded TC wire with the shield connected to analog common to reduce noise.
The USB-1616HS-2 several analog common screw terminals (see "Connecting the board for I/O operations"
starting on page 12). You can connect the shield of a shielded thermocouple to one of the analog commons.
When this connection is made, leave the shield at the other end of the thermocouple unconnected.
Caution! Connecting the shield to common at both ends results in a ground loop.
Averaging
Certain acquisition programs apply averaging after several samples have been collected. Depending on the
nature of the noise, averaging can reduce noise by the square root of the number of averaged samples.
22
USB-1616HS-2 User's Guide
Functional Details
Although averaging can be effective, it suffers from several drawbacks:
Noise in measurements only decreases as the square root of the number of measurements—reducing RMS
noise significantly may require many samples. Thus, averaging is suited to low-speed applications that can
provide many samples.
Only random noise is reduced or eliminated by averaging. Averaging does not reduce or eliminate periodic
signals.
Analog output
The USB-1616HS-2 has two 16-bit, 1 MHz analog output channels. Analog outputs can be updated at a
maximum rate of 1 MHz.
The channels have an output range of -10 V to +10 V. Each D/A can continuously output a waveform. In
addition, a program can asynchronously output a value to any of the D/A channels for non-waveform
applications, assuming that the D/A is not already being used in the waveform output mode.
When used to generate waveforms, you can clock the D/As in several different modes.
Internal output scan clock: The onboard programmable clock can generate updates ranging from 1 Hz to
1 MHz.
External output scan clock: A user-supplied external output scan clock at the DPR screw terminal.
External input scan clock: A user-supplied external input scan clock at APR can pace both the D/A and
the analog input.
Internal input scan clock: The internal ADC scan clock.
Example: Analog channel scanning of voltage inputs and streaming analog
outputs
The example shown in Figure 11 adds two DACs and a 16-bit digital pattern output to the example presented in
Figure 11. Analog channel scan of voltage inputs and streaming analog outputs example
This example updates all DACs and the 16-bits of digital I/O. These updates happen at the same time as the
output scan clock. All DACs and the 16-bits of pattern digital output are updated at the beginning of each scan.
23
USB-1616HS-2 User's Guide
Functional Details
Due to the time it takes to shift the digital data out to the DACs, plus the actual settling time of the digital-to-
analog conversion, the DACs actually take up to 4 µs after the start of scan to settle on the updated value.
The data for the DACs and pattern digital output comes from a PC-based buffer. The data is streamed across the
USB2 bus to the USB-1616HS-2.
You can update the DACs and pattern digital output with the output scan clock—either internally-generated or
externally-applied. In this scenario, the acquisition input scans are not synchronized to the analog outputs or
pattern digital outputs.
You can also synchronize everything—input scans, DACs, pattern digital outputs—to one clock, which is either
internally-generated or externally-applied.
Digital I/O
Twenty-four TTL-level digital I/O lines are included in each USB-1616HS-2. You can program digital I/O in
8-bit groups as either inputs or outputs and scan them in several modes (see "Digital input scanning" below).
You can access input ports asynchronously from the PC at any time, including when a scanned acquisition is
occurring.
Digital input scanning
Digital input ports can be read asynchronously before, during, or after an analog input scan.
Digital input ports can be part of the scan group and scanned along with analog input channels. Two
synchronous modes are supported when digital inputs are scanned along with analog inputs. Refer to "Example
4: Sampling digital inputs for every analog sample in a scan group" on page 13 for more information.
In both modes, adding digital input scans has no affect on the analog scan rate limitations.
If no analog inputs are being scanned, the digital inputs can sustain rates up to 4 MHz.
Higher rates—up to 12 MHz—are possible depending on the platform and the amount of data being transferred.
Digital outputs and pattern generation
Digital outputs can be updated asynchronously at anytime before, during, or after an acquisition. You can use
two of the 8-bit ports to generate a digital pattern at up to 4 MHz. The USB-1616HS-2 supports digital pattern
generation with bus mastering DMA. The digital pattern can be read from PC RAM.
Higher rates—up to 12 MHz—are possible depending on the platform and the amount of data being transferred.
Digital pattern generation is clocked using an internal clock. The onboard programmable clock generates
updates ranging from once every 1 second to 1 MHz, independent of any acquisition rate.
Triggering
Triggering can be the most critical aspect of a data acquisition application. The USB-1616HS-2 supports the
following trigger modes to accommodate certain measurement situations.
24
USB-1616HS-2 User's Guide
Functional Details
Hardware analog triggering
The USB-1616HS-2 uses true analog triggering in which the trigger level you program sets an analog DAC,
which is then compared in hardware to the analog input level on the selected channel. This guarantees an analog
trigger latency that is less than 1 µs.
You can select any analog channel as the trigger channel, but the selected channel must be the first channel in
the scan. You can program the trigger level, the rising or falling edge to trigger on, and hysteresis.
A note on the hardware analog level trigger and comparator change state
When analog input voltage starts near the trigger level, and you are performing a rising or falling hardware
analog level trigger, the analog level comparator may have already tripped before the sweep was enabled. If this
is the case, the circuit waits for the comparator to change state. However, since the comparator has already
changed state, the circuit does not see the transition.
To resolve this problem, do the following:
Set the analog level trigger to the threshold you want.
Apply an analog input signal that is more than 2.5% of the full-scale range away from the desired
threshold. This ensures that the comparator is in the proper state at the beginning of the acquisition.
Bring the analog input signal toward the desired threshold. When the input signal is at the threshold
(± some tolerance) the sweep will be triggered.
Before re-arming the trigger, again move the analog input signal to a level that is more than 2.5% of the
full-scale range away from the desired threshold.
For example, if you are using the ±2 V full-scale range (gain = 5), and you want to trigger at +1 V on the rising
edge, set the analog input voltage to a start value that is less than +0.9 V (1 V – (2 V * 2 * 2.5%)).
Digital triggering
A separate digital trigger input line is provided (TRG), allowing TTL-level triggering with latencies guaranteed
to be less than 1 µs. You can program both of the logic levels (1 or 0) and the rising or falling edge for the
discrete digital trigger input.
Software-based triggering
The three software-based trigger modes differ from hardware analog triggering and digital triggering because
the readings—analog, digital, or counter—are checked by the PC in order to detect the trigger event.
Analog triggering
You can select any analog channel as the trigger channel. You can program the trigger level, the rising or falling
edge to trigger on, and hysteresis.
Pattern triggering
You can select any scanned digital input channel pattern to trigger an acquisition, including the ability to mask
or ignore specific bits.
Counter triggering
You can program triggering to occur when one of the counters meets or exceeds a set value, or is within a range
of values. You can program any of the included counter channels as the trigger source.
25
USB-1616HS-2 User's Guide
Functional Details
Software-based triggering usually results in a long period of inactivity between the trigger condition being
detected and the data being acquired. However, the USB-1616HS-2 avoids this situation by using pre-trigger
data. When software-based-triggering is used, and the PC detects the trigger condition—which may be
thousands of readings after the actual occurrence of the signal—the USB-1616HS-2 driver automatically looks
back to the location in memory where the actual trigger-causing measurement occurred, and presents the
acquired data that begins at the point where the trigger-causing measurement occurs. The maximum inactive
period in this mode equals one scan period.
Stop trigger modes
You can use any of the software trigger modes explained previously to stop an acquisition.
For example, you can program an acquisition to begin on one event—such as a voltage level—and then stop on
another event—such as a digital pattern.
Pre-triggering and post-triggering modes
The USB-1616HS-2 supports four modes of pre-triggering and post-triggering, providing a wide-variety of
options to accommodate any measurement requirement.
When using pre-trigger, you must use software-based triggering to initiate an acquisition.
No pre-trigger, post-trigger stop event
In this simple mode, data acquisition starts when the trigger is received, and the acquisition stops when the stop-
trigger event is received.
Fixed pre-trigger with post-trigger stop event
In this mode, you set the number of pre-trigger readings to acquire. The acquisition continues until a stop-
trigger event occurs.
No pre-trigger, infinite post-trigger
In this mode, no pre-trigger data is acquired. Instead, data is acquired beginning with the trigger event, and is
terminated when you issue a command to halt the acquisition.
Fixed pre-trigger with infinite post-trigger
You set the amount of pre-trigger data to acquire. Then, the system continues to acquire data until the program
issues a command to halt acquisition.
Counter inputs
Four 32-bit counters are built into the USB-1616HS-2. Each counter accepts frequency inputs up to 20 MHz.
USB-1616HS-2 counter channels can be configured as standard counters or as multi-axis quadrature encoders.
The counters can concurrently monitor time periods, frequencies, pulses, and other event driven incremental
occurrences directly from pulse-generators, limit switches, proximity switches, and magnetic pick-ups.
Counter inputs can be read asynchronously under program control, or synchronously as part of an analog or
digital scan group.
26
USB-1616HS-2 User's Guide
Functional Details
When reading synchronously, all counters are set to zero at the start of an acquisition. When reading
asynchronously, counters may be cleared on each read, count up continually, or count until the 16-bit or 32-bit
limit has been reached. See counter mode explanations below.
Figure 12. Typical USB-1616HS-2 counter channel
Tips for making high-speed counter measurements (> 1 MHz)
Use coax or twisted-pair wire. Connect one side to Digital Common.
If the frequency source is tolerant, parallel-terminate the coax or twisted-pair with a 50 Ω or 100 Ω resistor
at the terminal block.
The amplitude of the driving waveform should be as high as possible without violating the over-voltage
specification.
To ensure adequate switching, waveforms should swing at least 0 V to 5 V and have a high slew rate.
Mapped channels
A mapped channel is one of four counter input signals that can get multiplexed into a counter module. The
mapped channel can participate with the counter's input signal by gating the counter, latching the counter, and
so on. The four possible choices for the mapped channel are the four counter input signals (post-debounce).
A mapped channel can be used to:
gate the counter
decrement the counter
latch the current count to the count register
Usually, all counter outputs are latched at the beginning of each scan within the acquisition. However, you can
use a second mapped channel to latch the counter output.
Counter modes
A counter can be asynchronously read with or without clear on read. The asynchronous read-signals strobe
when the lower 16-bits of the counter are read by software. The software can read the counter's high 16-bits
some time later after reading the lower 16-bits. The full 32-bit result reflects the timing of the first
asynchronous read strobe.
Totalize mode
The Totalize mode allows basic use of a 32-bit counter. While in this mode, the channel's input can only
increment the counter upward. When used as a 16-bit counter (counter low), one channel can be scanned at the
12 MHz rate. When used as a 32-bit counter (counter high), two sample times are used to return the full 32-bit
result. Therefore a 32-bit counter can only be sampled at a 6 MHz maximum rate. If you only want the upper
16 bits of a 32-bit counter, then you can acquire that upper word at the 12 MHz rate.
The counter counts up and does not clear on every new sample. However, it does clear at the start of a new scan
command.
27
USB-1616HS-2 User's Guide
Functional Details
The counter rolls over on the 16-bit (counter low) boundary, or on the 32-bit (counter high) boundary.
Clear on read mode
The counter counts up and is cleared after each read. By default, the counter counts up and only clears the
counter at the start of a new scan command. The final value of the counter—the value just before it was
cleared—is latched and returned to the USB-1616HS-2.
Clear on read mode is only available if the counter is in asynchronous mode the. The counter's lower 16-bit
value should be read first. This will latch the full 32-bit result and clear the counter. The upper 16-bit value can
be read after the lower 16-bit value.
Stop at the top mode
The counter stops at the top of its count. The top of the count is FFFF hex (65,535) for the 16-bit mode, and
FFFFFFFF hex (4,294,967,295) for the 32-bit mode.
32-bit or 16-bit
Sets the counter type to either 16-bits or 32-bits. The type of counter only matters if the counter is using the
stop at the top mode—otherwise, this option is ignored.
Latch on map
Sets the signal on the mapped counter input to latch the count.
By default, the start of scan signal—a signal internal to the USB-1616HS-2 that pulses once every scan period
to indicate the start of a scan group—latches the count so that the count is updated each time a scan is started.
Gating "on" mode
Sets the gating option to "on" for the mapped channel, enabling the mapped channel to gate the counter.
Any counter can be gated by the mapped channel. When the mapped channel is high, the counter is enabled.
When the mapped channel is low, the counter is disabled (but holds the count value). The mapped channel can
be any counter input channel other than the counter being gated.
Decrement "on" mode
Sets the counter decrement option to "on" for the mapped channel. The input channel for the counter
increments the counter, and you can use the mapped channel to decrement the counter.
Debounce modes
Each channel's output can be debounced with 16 programmable debounce times from 500 ns to 25.5 ms. The
debounce circuitry eliminates switch-induced transients typically associated with electro-mechanical devices
including relays, proximity switches, and encoders.
There are two debounce modes, as well as a debounce bypass, as shown in Figure 13. In addition, the signal
from the buffer can be inverted before it enters the debounce circuitry. The inverter is used to make the input
rising-edge or falling-edge sensitive.
Edge selection is available with or without debounce. In this case the debounce time setting is ignored and the
input signal goes straight from the inverter or inverter bypass to the counter module.
28
USB-1616HS-2 User's Guide
Functional Details
There are 16 different debounce times. In either debounce mode, the debounce time selected determines how
fast the signal can change and still be recognized.
The two debounce modes are trigger after stable and trigger before stable. A discussion of the two modes
follows.
Figure 13. Debounce model block diagram
Trigger after stable mode
In the trigger after stable mode, the output of the debounce module does not change state until a period of
stability has been achieved. This means that the input has an edge, and then must be stable for a period of time
equal to the debounce time.
Figure 14. Debounce module – trigger after stable mode
The following time periods (T1 through T5) pertain to Figure 14. In trigger after stable mode, the input signal
to the debounce module is required to have a period of stability after an incoming edge, in order for that edge to
be accepted (passed through to the counter module.) The debounce time for this example is equal to T2 and T5.
T1 – In the example above, the input signal goes high at the beginning of time period T1, but never stays
high for a period of time equal to the debounce time setting (equal to T2 for this example.)
T2 – At the end of time period T2, the input signal has transitioned high and stayed there for the required
amount of time—therefore the output transitions high. If the input signal does not stabilize in the high state
long enough, no transition would have appeared on the output and the entire disturbance on the input would
have been rejected.
T3 – During time period T3, the input signal remained steady. No change in output is seen.
T4 – During time period T4, the input signal has more disturbances and does not stabilize in any state long
enough. No change in the output is seen.
T5 – At the end of time period T5, the input signal has transitioned low and stayed there for the required
amount of time—therefore the output goes low.
Trigger before stable mode
In the trigger before stable mode, the output of the debounce module immediately changes state, but will not
change state again until a period of stability has passed. For this reason the mode can be used to detect glitches.
29
USB-1616HS-2 User's Guide
Functional Details
Figure 15. Debounce module – Trigger before stable mode
The following time periods (T1 through T6) pertain to the above drawing.
T1 – In the illustrated example, the input signal is low for the debounce time (equal to T1); therefore when
the input edge arrives at the end of time period T1, it is accepted and the output (of the debounce module)
goes high. Note that a period of stability must precede the edge in order for the edge to be accepted.
T2 – During time period T2, the input signal is not stable for a length of time equal to T1 (the debounce
time setting for this example.) Therefore, the output stays "high" and does not change state during time
period T2.
T3 – During time period T3, the input signal is stable for a time period equal to T1, meeting the debounce
requirement. The output is held at the high state. This is the same state as the input.
T4 – At anytime during time period T4, the input can change state. When this happens, the output will also
change state. At the end of time period T4, the input changes state, going low, and the output follows this
action [by going low].
T5 – During time period T5, the input signal again has disturbances that cause the input to not meet the
debounce time requirement. The output does not change state.
T6 – After time period T6, the input signal has been stable for the debounce time and therefore any edge on
the input after time period T6 is immediately reflected in the output of the debounce module.
Debounce mode comparisons
Figure 16 shows how the two modes interpret the same input signal, which exhibits glitches. Notice that the
trigger before stable mode recognizes more glitches than the trigger after stable mode. Use the bypass option to
achieve maximum glitch recognition.
Figure 16. Example of two debounce modes interpreting the same signal
Debounce times should be set according to the amount of instability expected in the input signal. Setting a
debounce time that is too short may result in unwanted glitches clocking the counter. Setting a debounce time
too long may result in an input signal being rejected entirely. Some experimentation may be required to find the
appropriate debounce time for a particular application.
To see the effects of different debounce time settings, simply view the analog waveform along with the counter
output. This can be done by connecting the source to an analog input.
30
USB-1616HS-2 User's Guide
Functional Details
Use trigger before stable mode when the input signal has groups of glitches and each group is to be counted as
one. The trigger before stable mode recognizes and counts the first glitch within a group but rejects the
subsequent glitches within the group if the debounce time is set accordingly. The debounce time should be set
to encompass one entire group of glitches as shown in the following diagram.
Figure 17. Optimal debounce time for trigger before stable mode
Trigger after stable mode behaves more like a traditional debounce function: rejecting glitches and only passing
state transitions after a required period of stability. Trigger after stable mode is used with electro-mechanical
devices like encoders and mechanical switches to reject switch bounce and disturbances due to a vibrating
encoder that is not otherwise moving. The debounce time should be set short enough to accept the desired input
Figure 18. Optimal debounce time for trigger after stable mode
Encoder mode
Rotary shaft encoders are frequently used with CNC equipment, metal-working machines, packaging
equipment, elevators, valve control systems, and in a multitude of other applications in which rotary shafts are
involved.
The USB-1616HS-2 supports quadrature encoders with up to 2 billion pulses per revolution, 20 MHz input
frequencies, and x1, x2, x4 count modes.
The encoder mode allows the USB-1616HS-2 to make use of data from optical incremental quadrature
encoders. In encoder mode, the USB-1616HS-2 accepts single-ended inputs. When reading phase A, phase B,
and index Z signals, the USB-1616HS-2 provides positioning, direction, and velocity data.
The USB-1616HS-2 can receive input from up to two encoders.
The USB-1616HS-2 supports quadrature encoders with a 16-bit (counter low) or a 32-bit (counter high)
counter, 20 MHz frequency, and X1, X2, and X4 count modes. With only phase A and phase B signals, two
channels are supported; with phase A, phase B, and index Z signals, 1 channel is supported. Each input can be
31
USB-1616HS-2 User's Guide
Functional Details
debounced from 500 ns to 25.5 ms (total of 16 selections) to eliminate extraneous noise or switch induced
transients. Encoder input signals must be within -5V to +10V and the switching threshold is TTL (1.3V).
Quadrature encoders generally have three outputs: A, B, and Z. The A and B signals are pulse trains driven by
an optical sensor inside the encoder. As the encoder shaft rotates, a laminated optical shield rotates inside the
encoder. The shield has three concentric circular patterns of alternating opaque and transparent windows
through which an LED shines. There is one LED and one phototransistor for each of the concentric circular
patterns. One phototransistor produces the A signal, another phototransistor produces the B signal and the last
phototransistor produces the Z signal. The concentric pattern for A has 512 window pairs (or 1024, 4096, etc.)
When using a counter for a trigger source, use a pre-trigger with a value of at least 1. Since all counters start at
zero with the initial scan, there is no valid reference in regard to rising or falling edge. Setting a pre-trigger to
1 or more ensures that a valid reference value is present, and that the first trigger is legitimate.
Figure 19. Representation of rotary shaft quadrature encoder
The concentric pattern for B has the same number of window pairs as A—except that the entire pattern is
rotated by 1/4 of a window-pair. Thus the B signal is always 90 degrees out of phase from the A signal. The A
and B signals pulse 512 times (or 1024, 4096, etc.) per complete rotation of the encoder.
The concentric pattern for the Z signal has only one transparent window and therefore pulses only once per
complete rotation. Representative signals are shown in the following figure.
A
B
Z
Figure 20. Representation of quadrature encoder outputs: A, B, and Z
As the encoder rotates, the A (or B) signal indicates the distance the encoder has traveled. The frequency of A
(or B) indicates the velocity of rotation of the encoder. If the Z signal is used to zero a counter (that is clocked
by A) then that counter will give the number of pulses the encoder has rotated from its reference. The Z signal is
a reference marker for the encoder. It should be noted that when the encoder is rotating clockwise (as viewed
from the back), A will lead B and when the encoder is rotating counterclockwise, A will lag B. If the counter
direction control logic is such that the counter counts upward when A leads B and counts downward when A
lags B, then the counter will give direction control as well as distance from the reference.
32
USB-1616HS-2 User's Guide
Functional Details
Maximizing encoder accuracy
If there are 512 pulses on A, then the encoder position is accurate to within 360°/512.
You can get even greater accuracy by counting not only rising edges on A but also falling edges on A, giving
position accuracy to 360 degrees/1024.
You get maximum accuracy counting rising and falling edges on A and on B (since B also has 512 pulses.) This
gives a position accuracy of 360°/2048. These different modes are known as X1, X2, and X4.
Connecting the USB-1616HS-2 to an encoder
You can use up to two encoders with each USB-1616HS-2 in your acquisition system. Each A and B signal can
be made as a single-ended connection with respect to common ground.
Differential applications are not supported.
For single-ended applications:
Connect signals A, B, and Z to the counter inputs on the USB-1616HS-2.
Connect each encoder ground to GND.
You can also connect external pull-up resistors to the USB-1616HS-2 counter input terminal blocks by placing
a pull-up resistor between any input channel and the encoder power supply. Choose a pull-up resistor value
based on the encoder's output drive capability and the input impedance of the USB-1616HS-2. Lower values of
pull-up resistors cause less distortion, but also cause the encoder's output driver to pull down with more current.
To external power
Ground (to Digital Common1)
Counter 0 (CNT0) – To Encoder “A”
Counter 1 (CNT1) – To Encoder “B”
To ground2
Counter 2 (CNT2) – To Encoder “Z”
1 The ground depicted at the left is associated with Digital Common on the USB-1616HS-4.
2 The ground depicted at the right is associated with the external power source.
Figure 21. Connections from single encoder to screw terminals on the USB-1616HS-2
The "A" signal must be connected to an even-numbered channel and the associated "B" signal must be
connected to the next higher odd-numbered channel. For example, if "A" were connected to counter 0, then "B"
would be connected to counter 1.
Connect each signal (A, B, Z) as a single-ended connection with respect to the common ground. The encoder
needs power from an external power output (typically +5 VDC). Connect the encoder's power input to the
power source and connect the return to the digital common of that source.
Wiring for two encoders: The following figure illustrates single-ended connections for two encoders.
Differential connections are not applicable.
33
USB-1616HS-2 User's Guide
Functional Details
Each signal (A, B) can be connected as a single-ended connection with respect to the common digital ground
(GND). Both encoders need power from an external power source (typically +5 VDC).
Connect each encoder's power input to the external power source. Connect the return to digital common (GND)
on the same source.
Ground (to Digital Common1)
To external power
To ground2
Counter 0 (CNT0) – To Encoder #1 “A”
Counter 1 (CNT1) – To Encoder #1 “B”
Counter 2 (CNT2) – To Encoder #2 “A”
Counter 3 (CNT3) – To Encoder #2 “B”
1 The ground depicted at the left is associated with Digital Common on the USB-1616HS-4.
2 The ground depicted at the right is associated with the external power source.
Figure 22. Connections from two encoders to screw terminals on the USB-1616HS-2
Timer outputs
Two 16-bit timer outputs are built into every USB-1616HS-2. Each timer output can generate a different square
wave with a programmable frequency in the range of 16 Hz to 1 MHz.
Figure 23. Typical USB-1616HS-2 timer channel
Example: Timer outputs
Timer outputs are programmable square waves. The period of the square wave can be as short as 1 µs or as long
as 65535 µs. Refer to the table below for examples of timer output frequencies.
Timer output frequency examples
Divisor
Timer output frequency
1
1 MHz
10 kHz
1 kHz
100
1000
10000
65535
100 Hz
15.259 Hz (in asynchronous write)
Turns timer off (for setpoint operation).
The two timer outputs can generate different square waves. The timer outputs can be updated asynchronously at
any time.
34
USB-1616HS-2 User's Guide
Functional Details
Both timer outputs can also be updated during an acquisition as the result of setpoints applied to analog or
digital inputs.
Using multiple USB-1616HS-2s per PC
USB-1616HS-2 features can be replicated up to four times, as up to four devices can be connected to a single
host PC. The serial number on each USB-1616HS-2 distinguishes one from another. You can operate multiple
USB-1616HS-2 boards synchronously. To do this, set up one USB-1616HS-2 with the pacer terminal you want
to use (APR or DPR) configured for output. Set up the USB-1616HS-2 boards you want to synchronize to this
board with the pacer screw terminal you want to use (APR or DPR) configured for input. Wire the pacer
terminal configured for output to each of the pacer input terminals that you want to synchronize.
To operate two or more USB-1616HS-2s synchronously:
Use coax (or twisted-pair wire) to connect the output signal to the input(s).
Connect Digital Common of each USB-1616HS-2 to one of the twisted pairs or to the shield of the coax.
Detection setpoint overview
You can program each as one of the following:
Single point referenced – Above, below, or equal to the defined setpoint.
Window (dual point) referenced – Inside or outside the window.
Window (dual point) referenced, hysteresis mode – Outside the window high forces one output (designated
Output 2; outside the window low-forces another output, designated as Output 1).
Figure 24. Diagram of detection setpoints
A digital detect signal is used to indicate when a signal condition is True or False—for example, whether or not
the signal has met the defined criteria. The detect signals can be part of the scan group and can be measured as
any other input channel, thus allowing real time data analysis during an acquisition.
The detection module looks at the 16-bit data being returned on a channel and generates another signal for each
channel with a setpoint applied (Detect1 for Channel 1, Detect2 for Channel 2, and so on). These signals serve
as data markers for each channel's data. It does not matter whether that data is volts, counts, or timing.
A channel's detect signal shows a rising edge and is True (1) when the channel's data meets the setpoint criteria.
The detect signal shows a falling edge and is False (0) when the channel's data does not meet the setpoint
criteria. The True and False states for each setpoint criteria are explained in the "Using the setpoint status
35
USB-1616HS-2 User's Guide
Functional Details
Criteria – input signal is equal to X
Action - driven by condition
Compare
X to:
Setpoint definition
(choose one)
Update conditions:
True only:
If True, then output value 1
If False, then perform no action
True and False:
If True, then output value 1
If False, then output value 2
Equal to A (X = A)
Below A (X < A)
Above B (X > B)
Limit A or
Limit B
True only
Window*
If True, then output value 1
(non-hyste Inside (B < X < A)
If False, then perform no action
resis
mode)
Outside ( B > X; or, X > A)
True and False
If True, then output value 1
If False, then output value 2
Hysteresis mode (forced update)
If X > A is True, then output value 2 until X < B is True, then
output value 1.
If X < B is True, then output value 1 until X > A is True, then
output value 2.
This is saying:
Above A (X > A)
Window*
(hysteresis
mode)
Below (A < X < B) (Both
conditions are checked when
in hysteresis mode
(a) If the input signal is outside the window high, output value 2 until
the signal goes outside the window low, and
(b) if the signal is outside the window low, output value 1 until the
signal goes outside the window high. There is no change to the detect
signal while within the window.
The detect signal has the timing resolution of the scan period as seen in the diagram below. The detect signal
can change no faster than the scan frequency (1/scan period.)
Figure 25. Example diagram of detection signals for channels 1, 2, and 3
Each channel in the scan group can have one detection setpoint. There can be no more than 16 total setpoints
total applied to channels within a scan group.
Detection setpoints act on 16-bit data only. Since the USB-1616HS-2 has 32-bit counters, data is returned
16-bits at a time. The lower word, the higher word, or both lower and higher words can be part of the scan
group. Each counter input channel can have one detection setpoint for the counter's lower 16-bit value and one
detection setpoint for the counter's higher 16-bit value.
36
USB-1616HS-2 User's Guide
Functional Details
Setpoint configuration
You program all setpoints as part of the pre-acquisition setup, similar to setting up an external trigger. Since
each setpoint acts on 16-bit data, each has two 16-bit compare values: a high limit (limit A) and a low limit
(limit B). These limits define the setpoint window.
There are several possible conditions (criteria) and effectively three update modes, as explained in the following
configuration summary.
Set high limit
You can set the 16-bit high limit (limit A) when configuring the USB-1616HS-2 through software.
Set low limit
You can set the 16-bit low limit (limit B) when configuring the USB-1616HS-2 through software.
Set criteria
Inside window: Signal is below 16-bit high limit and above 16-bit low limit.
Outside window: Signal is above 16-bit high limit, or below 16-bit low limit.
Greater than value: Signal is above 16-bit low limit, so 16-bit high limit is not used.
Less than value: Signal is below 16-bit high limit, so 16-bit low limit is not used.
Equal to value: Signal is equal to 16-bit high limit, and limit B is not used.
The equal to mode is intended for use when the counter or digital input channels are the source channel.
You should only use the equal to16-bit high limit (limit A) mode with counter or digital input channels as
the channel source. If you want similar functionality for analog channels, then use the inside window mode
Hysteresis mode: Outside the window, high forces output 2 until an outside the window low condition
exists, then output 1 is forced. Output 1 continues until an outside the window high condition exists. The
cycle repeats as long as the acquisition is running in hysteresis mode.
Set output channel
None
Update FIRSTPORTC
Update DAC
Update timerx
Update modes
Update on True only
Update on True and False
Set values for output
16-bit DAC value, FIRSTPORTC* value, or timer value when input meets criteria.
16-bit DAC value, FIRSTPORTC* value, or timer value when does not meet criteria.
* By default, FIRSTPORTC comes up as a digital input. You may want to initialize FIRSTPORTC to a
known state before running the input scan to detect the setpoints.
When using setpoints with triggers other than immediate, hardware analog, or TLL, the setpoint criteria
evaluation begins immediately upon arming the acquisition.
Using the setpoint status register
You can use the setpoint status register to check the current state of the 16 possible setpoints. In the register,
Setpoint 0 is the least-significant bit and Setpoint 15 is the most-significant bit. Each setpoint is assigned a
value of 0 or 1.
37
USB-1616HS-2 User's Guide
Functional Details
A value of 0 indicates that the setpoint criteria are not met—in other words, the condition is False.
A value of 1 indicates that the criteria have been met—in other words, the condition is True.
In the following example, the criteria for setpoints 0, 1, and 4 is satisfied (True), but the criteria for the other
13 setpoints has not been met.
Setpoint #
True (1)
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0
0
0
0
0
0
0
0 0 0 0 0 1 0 0 1 1
<<< Most significant bit
Least significant bit >>>
False (0)
From the table above we have 10011binary, or 19 decimal, derived as follows:
Setpoint 0, having a True state, shows 1, giving us decimal 1.
Setpoint 1, having a True state, shows 1, giving us decimal 2.
Setpoint 4, having a True state, shows 1, giving us decimal 16.
For proper operation, the setpoint status register must be the last channel in the scan list.
Examples of control outputs
Detecting on analog input, DAC, and FIRSTPORTC updates
Update mode: Update on True and False
Criteria: Channel 5 example: below limit; channel 4 example: inside window
In this example, channel 5 is programmed with reference to one setpoint (limit A), defining a low limit.
Channel 4 is programmed with reference to two setpoints (limit A and limit B) which define a window for that
channel.
Channel
Condition
State of detect signal Action
5
Below limit A (for
channel 5)
True
False
True
False
When channel 5 analog input voltage is below the limit
A, update DAC1 with output value 0.0 V.
When the above stated condition is false, update DAC1
with the Output Value of - 1.0 V.
4
Within window
(between limit A and
limit B) for channel 4
When Channel 4's analog input voltage is within the
window, update FIRSTPORTC with 70h.
When the above stated condition is False (channel 4
analog input voltage is outside the window), update
FIRSTPORTC with 30h.
38
USB-1616HS-2 User's Guide
Functional Details
Figure 26. Analog inputs with setpoints update on True and False
In the channel 5 example, the setpoint placed on analog Channel 5 updated DAC1 with 0.0 V. The update
occurred when channel 5's input was less than the setpoint (limit A). When the value of channel 5's input was
above setpoint limit A, the condition of <A was false and DAC1 was then updated with -1.0 V.
You can program control outputs programmed on each setpoint, and use the detection for channel 4 to update
the FIRSTPORTC digital output port with one value (70 h in the example) when the analog input voltage is
within the shaded region and a different value when the analog input voltage is outside the shaded region (30 h
in the example).
Detection on an analog input, timer output updates
Update Mode: Update on True and False
Criteria Used: Inside window
Figure 27 shows how a setpoint can be used to update a timer output. Channel 3 is an analog input channel. A
setpoint is applied using update on True and False, with a criteria of inside-the-window, where the signal value
is inside the window when simultaneously less than Limit A but greater than Limit B.
Whenever the channel 3 analog input voltage is inside the setpoint window (condition True), Timer0 is updated
with one value; and whenever the channel 3 analog input voltage is outside the setpoint window (condition
False) timer0 will be updated with a second output value.
39
USB-1616HS-2 User's Guide
Functional Details
Figure 27. Timer output update on True and False
Using the hysteresis function
Update mode: N/A, the hysteresis option has a forced update built into the function
Criteria used: Window criteria for above and below the set limits
Figure 28 shows analog input Channel 3 with a setpoint which defines two 16-bit limits, Limit A (High) and
Limit B (Low). These are being applied in the hysteresis mode and DAC channel 0 is updated accordingly.
In this example, Channel 3's analog input voltage is being used to update DAC0 as follows:
When outside the window, low (below limit B) DAC0 is updated with 3.0 V. This update remains in effect
until the analog input voltage goes above Limit A.
When outside the window, high (above limit A), DAC0 is updated with 7.0 V. This update remains in
effect until the analog input signal falls below limit B. At that time we are again outside the limit "low" and
the update process repeats itself.
Hysteresis mode can also be done with FIRSTPORTC digital output port, or a timer output, instead of a DAC.
Figure 28. Channel 3 in hysteresis mode
40
USB-1616HS-2 User's Guide
Functional Details
Using multiple inputs to control one DAC output
Update mode: Rising edge, for each of two channels
Criteria used: Inside window, for each of two channels
The figure below shows how multiple inputs can update one output. In the following figure the DAC2 analog
output is being updated. Analog input Channel 3 has an inside-the-window setpoint applied. Whenever Channel
3's input goes inside the programmed window, DAC2 will be updated with 3.0 V.
Analog input Channel 7 also has an inside-the-window setpoint applied. Whenever channel 7's input goes inside
the programmed window, DAC2 is updated with - 7.0 V.
Figure 29. Using two criteria to control an output*
The update on True-only mode was selected, and therefore the updates for DAC2 only occur when the criteria is
met. However, in the above figure we see that there are two setpoints acting on one DAC. We can also see that
the two criteria can be met simultaneously. When both criteria are True at the same time, the DAC2 voltage is
associated with the criteria that has been most recently met.
Detecting setpoints on a totalizing counter
In the following figure, Channel 1 is a counter in totalize mode. Two setpoints define a point of change for
Detect 1 as the counter counts upward. The detect output is high when inside the window (greater than Limit B
(the low limit) but less than Limit A (the high limit).
In this case, the Channel 1 setpoint is defined for the 16 lower bits of channel 1's 32-bit value. The
FIRSTPORTC digital output port could be updated on a True condition (the rising edge of the detection signal).
Alternately, one of the DAC output channels, or timer outputs, could be updated with a value.
41
USB-1616HS-2 User's Guide
Functional Details
At this point you can update FIRSTPORTC or DACs
Figure 30. Channel 1 in totalizing counter mode, inside the window setpoint
Detection setpoint details
Controlling analog, digital, and timer outputs
You can program each setpoint with an 8-bit digital output byte and corresponding 8-bit mask byte. When the
setpoint criteria is met, the FIRSTPORTC digital output port can be updated with the given byte and mask.
Alternately, you can program each setpoint with a 16-bit DAC update value, and any one of the two DAC
outputs can be updated in real time. Any setpoint can also be programmed with a timer update value.
In hysteresis mode, each setpoint has two forced update values. Each update value can drive one DAC, one
timer, or the FIRSTPORTC digital output port. In hysteresis mode, the outputs do not change when the input
values are inside the window. There is one update value that gets applied when the input values are less than the
window and a different update value that gets applied when the input values are greater than the window.
Update on True and False uses two update values. The update values can drive DACs, FIRSTPORTC, or timer
outputs.
FIRSTPORTC digital outputs can be updated immediately upon setpoint detection. This is not the case for
analog outputs, as these incur another 3 µs delay. This is due to the shifting of the digital data out to the D/A
converter which takes 1µs, plus the actual conversion time of the D/A converter, i.e., another 2µs (worst case).
Going back to the above example, if the setpoint for analog input Channel 2 required a DAC update it would
occur 5µs after the ADC conversion for Channel 2, or 6µs after the start of the scan.
When using setpoints to control any of the DAC outputs, increased latencies may occur if attempting to stream
data to DACs or pattern digital output at the same time. The increased latency can be as long as the period of
the output scan clock. For these reasons, avoid streaming outputs on any DAC or pattern digital output when
using setpoints to control DACs.
FIRSTPORTC, DAC, or timer update latency
Setpoints allow analog outputs, DACs, timers, or FIRSTPORTC digital outputs to update very quickly. Exactly
how fast an output can update is determined by these factors:
scan rate
synchronous sampling mode
type of output to be updated
42
USB-1616HS-2 User's Guide
Functional Details
For example, you set an acquisition to have a scan rate of 100 kHz, which means each scan period is 10 µs.
Within the scan period you sample six analog input channels. These are shown in the following figure as
channels 1 through 6. The ADC conversion occurs at the beginning of each channel's 1 µs time block.
FIRSTPORTC
Figure 31. Example of FIRSTPORTC or DAC latency
By applying a setpoint on analog input channel 2, that setpoint gets evaluated every 10 µs with respect to the
sampled data for channel 2.
Due to the pipelined architecture of the analog-to-digital converter system, the setpoint cannot be evaluated
until 2 µs after the ADC conversion. In the example above, the FIRSTPORTC digital output port can be
updated no sooner than 2 µs after channel 2 has been sampled, or 3 µs after the start of the scan. This 2 µs delay
is due to the pipelined ADC architecture. The setpoint is evaluated 2 µs after the ADC conversion and then
FIRSTPORTC can be updated immediately.
The detection circuit works on data that is put into the acquisition stream at the scan rate. This data is acquired
according to the pre-acquisition setup (scan group, scan period, etc.) and returned to the PC. Counters are
latched into the acquisition stream at the beginning of every scan. The actual counters may be counting much
faster than the scan rate, and therefore only every 10th, 100th, or nth count shows up in the acquisition data.
As a result, you can set a small detection window on a totalizing counter channel and have the detection setpoint
"stepped over" since the scan period was too long. Even though the counter value stepped into and out of the
detection window, the actual values going back to the PC may not. This is true no matter what mode the counter
channel is in.
When setting a detection window, keep a scan period in mind. This applies to analog inputs and counter inputs.
Quickly changing analog input voltages can step over a setpoint window if not sampled often enough.
There are three possible solutions for overcoming this problem:
Shorten the scan period to give more timing resolution on the counter values or analog values.
Widen the setpoint window by increasing limit A and/or lowering limit B.
A combination of both solutions (1 and 2) could be made.
43
Chapter 4
Calibrating the USB-1616HS-2
Every range of a USB-1616HS-2 device is calibrated at the factory using a digital NIST traceable calibration
method. This method works by storing a correction factor for each range on the unit at the time of calibration.
For analog inputs, the user can adjust the calibration of the board while it is installed in the acquisition system
without destroying the factory calibration supplied with the board. This is accomplished by having two distinct
calibration tables in the USB-1616HS-2 on-board EPROM—one which contains the factory calibration, and the
other which is available for field calibration.
You can perform field calibration automatically in seconds with InstaCal and without the use of external
hardware or instruments.
Field calibration derives its traceability through an on-board reference which has a stability of 0.005% per year.
Note that a two-year calibration period is recommended for USB-1616HS-2 boards. You should calibrate the
USB-1616HS-2 using InstaCal after the board has fully warmed up. The recommended warm-up time is 30
minutes. For best results, calibrate the board immediately before making critical measurements. The high
resolution analog components on the board are somewhat sensitive to temperature. Pre-measurement calibration
ensures that your board is operating at optimum calibration values.
44
Chapter 5
Specifications
Typical for 25 °C unless otherwise specified.
Specifications in italic text are guaranteed by design.
Analog input
Table 1. Analog input specifications
Successive approximation
A/D converter type
Resolution
16 bits
Number of channels
16 single-ended/8 differential, software-selectable.
Up to 48 additional analog inputs per module are available with the optional AI-EXP48
module. Expansion channel features are the same as those of the main channels.
Input ranges (software and
sequencer programmable)
Bipolar: ±10 V, ±5 V, ±2 V, ±1 V , ±0.5 V, ±0.2 V, ±0.1 V
Maximum sample rate
Nonlinearity (integral)
Nonlinearity (differential)
A/D pacing
1 MHz
±2 LSB maximum
±1 LSB maximum
Onboard input scan clock, external source (APR)
Trigger sources and modes
Acquisition data buffer
See Table 8
1 MSample
DMA
Data transfer
Configuration memory
Programmable I/O
Maximum usable input
voltage + common mode
voltage (CMV + Vin)
Range: ±10 V, ±5 V, ±2 V, ±1 V, ±0.5 V
10.5 V maximum
2.1 V maximum
Range: ±0.2 V, ±0.1 V
Signal to noise and distortion
Total harmonic distortion
Calibration
72 dB typical for ±10 V range, 1 kHz fundamental
-80 dB typical for ±10 V range, 1 kHz fundamental
Auto-calibration, calibration factors for each range stored onboard in non-volatile RAM.
-70 dB typical DC to 1 kHz
CMRR @ 60 Hz
Bias current
40 pA typical (0 °C to 35°C)
Crosstalk
-75 dB typical DC to 60 Hz; -65 dB typical @ 10 kHz
10 MΩ single-ended, 20 MΩ differential
±30 V
Input impedance
Absolute maximum input
voltage
Accuracy
Table 2. Analog input accuracy specifications
Voltage range
Accuracy
±(% of reading + %
range)
Temperature coefficient
±(ppm of reading + ppm range)/°C
Noise (cts
RMS)
23°C ±10 °C, 1 year
-10 V to 10 V
-5 V to 5 V
-2 V to 2 V
0.031% + 0.008%
0.031% + 0.009%
0.031% + 0.010%
0.031% + 0.02%
0.031% + 0.04%
0.036% + 0.075%
0.042% + 0.15%
14 + 8
14 + 9
14 +10
14 + 12
14 +18
14 +12
14 +18
2.0
3.0
2.0
Note 1
Note 2
-1 V to 1 V
3.5
5.5
8.0
14.0
-500 mV to 500 mV
-200 mV to 200 mV
-100 mV to 100 mV
45
USB-1616HS-2 User's Guide
Specifications
Note 1: Specifications assume differential input single-channel scan, 1 MHz scan rate, unfiltered,
CMV=0.0 V, 30 minute warm-up, exclusive of noise, range is +FS to -FS.
Note 2: Noise reflects 10,000 samples at 1 MHz, typical, differential short.
Thermocouples
Table 3. Thermocouple (TC) types and accuracy (Note 3)
TC type
Temperature range (°C)
Accuracy (±°C)
Noise typical (±°C)
J
-200 to + 760
-200 to + 1200
-200 to + 400
-270 to + 650
-50 to + 1768
-50 to + 1768
-270 to + 1300
+300 to + 1400
1.7
1.8
1.8
1.7
4.8
4.7
2.7
3.0
0.2
0.2
0.2
0.2
1.5
1.5
0.3
1.0
K
T
E
R
S
N
B
Note 3: Assumes 16384 oversampling applied, CMV = 0.0V, 60 minute warm-up, still environment, and
25 °C ambient temperature; excludes thermocouple error; TCin = 0° C for all types except B
(1000 °C), TR-2U power supply for external power.
Analog outputs
Analog output channels can be updated synchronously relative to scanned inputs, and clocked from either an
internal onboard clock, or an external clock source. Analog outputs can also be updated asynchronously,
independent of any other scanning system.
Table 4. Analog output specifications
Channels
2
Resolution
16-bits
Data buffer
PC-based memory
Output voltage range
Output current
±10 V
±1 mA
Sourcing more current (1 to 10 mA) may require a TR-2U power supply.
Offset error
±0.0045 V maximum
<10 mV when updated
<12 mV typical at major carry
±0.01%
Digital feed-through
DAC analog glitch
Gain error
Coupling
DC
Update rate
1 MHz maximum, resolution 20.83 ns
2 µs to rated accuracy
Settling time
Pacer sources
Four programmable sources:
Onboard output scan clock, independent of input scan clock
Onboard input scan clock
External output scan clock (DPR), independent of external input scan clock (APR)
External input scan clock (APR)
Trigger sources
Start of input scan
46
USB-1616HS-2 User's Guide
Specifications
Digital input/output
Table 5. Digital input/output specifications
Number of I/O
Ports
24
Three banks of eight.
Each port is programmable as input or output
Input scanning modes
Two programmable
Asynchronous, under program control at any time relative to input scanning
Synchronous with input scanning
Input characteristics
Logic keeper circuit
Input protection
Input high
220 Ω series resistors, 20 pF to common
Holds the logic value to 0 or 1 when there is no external driver
±15 kV ESD clamp diodes parallel
+2.0 V to +5.0 V
Input low
0 to 0.8 V
Output high
>2.0 V
Output low
<0.8 V
Output current
Output 1.0 mA per pin
Sourcing more current may require a TR-2U power supply.
Onboard input scan clock, external input scan clock (APR)
Digital input pacing
Digital output pacing
Four programmable sources:
Onboard output scan clock, independent of input scan clock
Onboard input scan clock
External output scan clock (DPR), independent of external input scan clock
(APR)
External input scan clock (APR)
Digital input trigger sources and
modes
Digital output trigger sources
Data transfer
Start of input scan
DMA
Sampling/update rate
Pattern generation output
4 MHz maximum (rates up to 12 MHz are sustainable on some platforms)
Two of the 8-bit ports can be configured for 16-bit pattern generation. The pattern
can also be updated synchronously with an acquisition at up to 4 MHz.
47
USB-1616HS-2 User's Guide
Specifications
Counters
Counter inputs can be scanned based on an internal programmable timer or an external clock source.
Table 6. Counter specifications
Channels
4 independent
Resolution
32-bit
Input frequency
Input signal range
Input characteristics
Trigger level
20 MHz maximum
-5 V to 10 V
10 kΩ pull-up, 200 Ω series resistor, ±15 kV ESD protection
TTL
Minimum pulse width
Debounce times
25 ns high, 25 ns low
16 selections from 500 ns to 25.5 ms, positive or negative edge sensitive, glitch
detect mode or debounce mode
Time base accuracy
Counter read pacer
50 ppm (0 ° to 50 °C)
Onboard input scan clock, external input scan clock (APR)
Trigger sources and modes
Programmable mode
Counter mode options
Counter
Totalize, clear on read, rollover, stop at all Fs, 16- or 32-bit, any other channel can
gate the counter
Input sequencer
Analog, digital, and counter inputs can be scanned based on either an internal programmable timer or an
external clock source.
Table 7. Input sequencer specifications
Input scan clock sources: two (see Note 4) Internal, programmable:
Analog channels from 1 µs to 1 s in 20.83 ns steps.
Digital channels and counters from 250 ns to 1 s in 20.83 ns steps.
External. TTL level input (APR):
Analog channels down to 1 µs minimum
Digital channels and counters down to 250 ns minimum
Programmable parameters per scan:
Depth
Programmable channels (random order), programmable gain
512 locations
Onboard channel-to-channel scan rate
Analog: 1 MHz maximum
Digital: 4 MHz if no analog channels are enabled, 1 MHz with analog
channels enabled
External input scan clock (APR) maximum Analog: 1.0 MHz
rate
Digital: 4 MHz if no analog channels are enabled, 1 MHz with analog
channels enabled
Clock signal range:
Logical zero: 0 V to 0.8 V
Logical one: 2.4 V to 5.0 V
50 ns high, 50 ns low
Minimum pulse width
Note 4: The maximum scan clock rate is the inverse of the minimum scan period. The minimum scan
period is equal to 1 µs times the number of analog channels. If a scan contains only digital
channels, then the minimum scan period is 250 ns.
Some platforms can sustain clock rates up to 83.33 ns.
48
USB-1616HS-2 User's Guide
Triggering
Specifications
Table 8. Trigger sources and modes
Explanation
Trigger source
Single channel analog Any analog input channel can be software programmed as
hardware trigger
the analog trigger channel, including any of the analog expansion channels.
Input signal range: -10 V to +10 V maximum
Trigger level: Programmable (12-bit resolution)
Latency: 350 ns typical, 1.3 µs max
Accuracy: ±0.5% of reading, ±2 mV offset maximum
Noise: 2 mV RMS typical
Single channel analog
software trigger
Any analog input channel—including any of the analog expansion channels, can be selected as
the software trigger channel. If the trigger channel involves a calculation, such as temperature,
then the driver automatically compensates for the delay required to obtain the reading, resulting
in a maximum latency of one scan period.
Input signal range: Anywhere within range of the trigger channel
Trigger level: Programmable (16-bit resolution)
Latency: One scan period (maximum)
External-single
channel digital trigger
A separate digital input is provided for digital triggering.
Input signal range: -15 V to +15 V maximum
Trigger level: TTL level sensitive
Minimum pulse width: 50 ns high, 50 ns low
Latency: One scan period maximum
Digital pattern
triggering
8-bit or 16-bit pattern triggering on any of the digital ports. Programmable for trigger on equal,
not equal, above, or below a value.
Individual bits can be masked for "don't care" condition.
Latency: One scan period, maximum
Counter/totalizer
triggering
Counter/totalizer inputs can trigger an acquisition. User can select to trigger on a frequency or on
total counts that are equal, not equal, above, or below a value, or within/outside of a window
rising/falling edge.
Latency: One scan period, maximum
Frequency/pulse generators
Table 9. Frequency/pulse generator specifications
Channels
2 × 16-bit
Output waveform
Output rate
Square wave
1 MHz base rate divided by 1 to 65535 (programmable)
2.0 V minimum @ -1.0 mA, 2.9 V minimum @ -400 µA
0.4 V maximum @ 400 µA
High-level output voltage
Low-level output voltage
Power consumption
Power consumption specification is for a USB-1616HS-2. Add 400mW for a USB-1616HS-2 connected to an
AI-EXP48 expansion module.
Table 10. Power consumption specifications (Note 5)
Power consumption (per board) 3000 mW
49
USB-1616HS-2 User's Guide
Specifications
External power
Table 11. External power specifications (Note 5)
Connector
Switchcraft # RAPC-712
6 to 16 VDC (used when USB port supplies insufficient power, or when an independent
power supply is desired)
Power range
20 V for 10 seconds, maximum
Over-voltage
Note 5: The power supply (MCC p/n TR-2U) and line cord (MCC p/n CA-1) are required if the USB port
cannot supply adequate power. By USB 2.0 standards, USB 2.0 ports must supply 2500 mW
(nominal at 5 V, 500 mA)
USB specifications
Table 12. USB specifications
USB-device type
USB 2.0 high-speed mode (480 Mbps) if available (recommended), otherwise, USB1.1
full-speed mode (12 Mbps)
USB 2.0 (recommended) or USB 1.1
Device compatibility
Environmental
Table 13. Environmental specifications
Operating temperature range
Storage temperature range
Relative humidity
-30 °C to +70 °C
-40 °C to +80 °C
0 to 95% non-condensing
Mechanical
Table 14. Mechanical specifications
Vibration
Dimensions
Weight
MIL STD 810E category 1 and 10
269 mm (W) x 92 mm (D) x 45 mm (H )(10.6” x 3.6” x 1.6”)
431 g (0.95 lbs)
Signal I/O connectors and pin out
Table 15. Screw connector specifications
Connector type
Screw terminal
Wire gauge range
Expansion connector type
14 AWG to 30 AWG
25-pin DSUB, female
Compatible expansion products AI-EXP48 expansion module with screw terminals
50
USB-1616HS-2 User's Guide
Specifications
Table 16. USB-1616HS-2 screw terminal pin out – single-ended connections
Analog common (Ab)
Analog output 0 (AO0)
Analog output 1 (AO1)
NC
Digital common (Db)
FIRSTPORTA Bit 0 (A0)
FIRSTPORTA Bit 1 (A1)
FIRSTPORTA Bit 2 (A2)
FIRSTPORTA Bit 3 (A3)
FIRSTPORTA Bit 4 (A4)
FIRSTPORTA Bit 5 (A5)
FIRSTPORTA Bit 6 (A6)
FIRSTPORTA Bit 7 (A7)
Digital common (Db)
Timer 0 (T0)
NC
Analog common (Ab)
CAL (Reserved for self-calibration)
Signal ground (Sb)
Digital common (Db)
TTL trigger (TRG)
Analog Out
DIG-Tmr I/O
Output scan clock I/O (DPR)
Input scan clock I/O (APR)
Timer 1 (T1)
Analog common (Ab)
CH 0 (0H)
CH 8 (8L)
Analog common (Ab)
CH 1 (1H)
CH 9 (9L)
Analog common (Ab)
CH 2 (2H)
CH 10 (10L)
Analog common (Ab)
CH 3 (3H)
Digital common (Db)
FIRSTPORTB Bit 0 (B0)
FIRSTPORTB Bit 1 (B1)
FIRSTPORTB Bit 2 (B2)
FIRSTPORTB Bit 3 (B3)
FIRSTPORTB Bit 4 (B4)
FIRSTPORTB Bit 5 (B5)
FIRSTPORTB Bit 6 (B6)
FIRSTPORTB Bit 7 (B7)
Digital common (Db)
Analog In
Dig-Ctr I/O
Counter 0 (CT0)
Counter 1 (CT1)
CH 11 (11L)
Analog common (Ab)
CH 4 (4H)
CH 12 (12L)
Analog common (Ab)
CH 5 (5H)
CH 13 (13L)
Analog common (Ab)
CH 6 (6H)
CH 14 (14L)
Analog common (Ab)
CH 7 (7H)
Digital common (Db)
FIRSTPORTC Bit 0 (C0)
FIRSTPORTC Bit 1 (C1)
FIRSTPORTC Bit 2 (C2)
FIRSTPORTC Bit 3 (C3)
FIRSTPORTC Bit 4 (C4)
FIRSTPORTC Bit 5 (C5)
FIRSTPORTC Bit 6 (C6)
FIRSTPORTC Bit 7 (C7)
Digital common (Db)
Counter 2 (CT2)
Analog In
Dig-Ctr I/O
CH 15 (15L)
Counter 3 (CT3)
51
USB-1616HS-2 User's Guide
Specifications
Table 17. USB-1616HS-2 screw terminal pin out – differential connections
Analog common (Ab)
Analog output 0 (AO0)
Analog output 1 (AO1)
NC
Digital common (Db)
FIRSTPORTA Bit 0 (A0)
FIRSTPORTA Bit 1 (A1)
FIRSTPORTA Bit 2 (A2)
FIRSTPORTA Bit 3 (A3)
FIRSTPORTA Bit 4 (A4)
FIRSTPORTA Bit 5 (A5)
FIRSTPORTA Bit 6 (A6)
FIRSTPORTA Bit 7 (A7)
Digital common (Db)
Timer 0 (T0)
NC
Analog common (Ab)
CAL (Reserved for self-calibration)
Signal ground (Sb)
Digital common (Db)
TTL trigger (TRG)
Analog Out
DIG-Tmr I/O
Output scan clock I/O (DPR)
Input scan clock I/O (APR)
Timer 1 (T1)
Analog common (Ab)
CH 0 HI (0H)
CH 0 LO (8L)
Analog common (Ab)
CH 1 HI (1H)
CH 1 LO (9L)
Analog common (Ab)
CH 2 HI (2H)
CH 2 LO (10L)
Analog common (Ab)
CH 3 HI (3H)
Digital common (Db)
FIRSTPORTB Bit 0 (B0)
FIRSTPORTB Bit 1 (B1)
FIRSTPORTB Bit 2 (B2)
FIRSTPORTB Bit 3 (B3)
FIRSTPORTB Bit 4 (B4)
FIRSTPORTB Bit 5 (B5)
FIRSTPORTB Bit 6 (B6)
FIRSTPORTB Bit 7 (B7)
Digital common (Db)
Analog In
Dig-Ctr I/O
Counter 0 (CT0)
Counter 1 (CT1)
CH 3 LO (11L)
Analog common (Ab)
CH 4 HI (4H)
CH 4 LO (12L)
Analog common (Ab)
CH 5 HI (5H)
CH 5 LO (13L)
Analog common (Ab)
CH 6 HI (6H)
CH 6 LO (14L)
Analog common (Ab)
CH 7 HI (7H)
Digital common (Db)
FIRSTPORTC Bit 0 (C0)
FIRSTPORTC Bit 1 (C1)
FIRSTPORTC Bit 2 (C2)
FIRSTPORTC Bit 3 (C3)
FIRSTPORTC Bit 4 (C4)
FIRSTPORTC Bit 5 (C5)
FIRSTPORTC Bit 6 (C6)
FIRSTPORTC Bit 7 (C7)
Digital common (Db)
Counter 2 (CT2)
Analog In
Dig-Ctr I/O
CH 7 LO (15L)
Counter 3 (CT3)
52
Declaration of Conformity
Manufacturer:
Address:
IOTech, Incorporated
25971 Cannon Road
Cleveland, OH 44146
USA
Category:
Information technology equipment.
IOTech, Incorporated declares under sole responsibility that the product
USB-1616HS-2
to which this declaration relates is in conformity with the relevant provisions of the following standards or other
documents:
EU EMC Directive 89/336/EEC: Electromagnetic Compatibility, EN 61326 (1997) Amendment 1 (1998)
Emissions: Group 1, Class A
EN 55022 (1993)/CISPR 22: Radiated and Conducted emissions.
Immunity: EN61326, Annex A
IEC 61000-4-2 (1995): Electrostatic Discharge immunity, Criteria B.
IEC 61000-4-3 (1995): Radiated Electromagnetic Field immunity Criteria A.
IEC 61000-4-4 (1995): Electric Fast Transient Burst immunity Criteria A.
IEC 61000-4-6 (1996): Radio Frequency Common Mode immunity Criteria A.
IEC 61000-4-11 (1994): Voltage Dips, Interruption immunity.
To maintain the safety, emission, and immunity standards of this declaration, the following conditions must be
met.
The host computer, peripheral equipment, power sources, and expansion hardware must be CE compliant.
Equipment must be operated in a controlled electromagnetic environment as defined by Standards EN
61326:1998, or IEC 61326:1998.
Shielded wires must be used for all I/Os and must be less than 3 meters (9.75 feet) in length. Clips must be used with
the AI-EXP48.
The host computer must be properly grounded.
The host computer must be USB2.0 compliant and IOtech USB cables (CA-179-x) must be used.
If using the USB-1616HS Series device in a high RF environment (3 to 10 V/m), then a clamp-on ferrite (IOtech p/n L-
8-1) may be needed on the USB cable, otherwise communication may be disrupted.
A protective ESD wrist strap should be used when connecting or disconnecting leads from screw terminal blocks.
Alternatively, unplug the unit from the host computer when making connections. Protective housings (IOtech p/n
CN-241-12) can be placed over the removable terminal blocks to protect signals from ESD during operation.
If external DC power is needed, a TR-2U power supply must be used.
Note: Data acquisition equipment may exhibit noise or increased offsets when exposed to high RF fields
(>3V/m) or transients.
Declaration of Conformity based on tests conducted by Smith Electronics, Inc., Cleveland, OH 44141, USA in
December, 2005. Test records are outlined in Smith Electronics Test Report “Personal Daq/3000 Series with
PDQ30 Expansion Module” and “PDAQ3000-PDQ30 Addenda”.
We hereby declare that the equipment specified conforms to the above Directives and Standards.
Paul Wittibschlager
Director of Hardware Engineering
Measurement Computing Corporation