Agilent Technologies Agilent B2201A User Manual

Agilent B2200A  
Femto Leakage Switch  
Mainframe  
Agilent B2201A  
14ch Low Leakage Switch  
Mainframe  
User’s Guide  
Agilent Technologies  
DECLARATION OF CONFORMITY  
According to ISO/IEC Guide 22 and CEN/CENELEC EN 45014  
Manufacturer’s Name :  
Manufacturers Address:  
Supplier’s Address :  
Agilent Technologies International sarl  
Rue de la Gare 29  
CH - 1110 Morge s  
Switzerland  
Declares under sole responsibility that the product as originally delivere d  
Product Name :  
Model Number :  
Product Options:  
Femto Leakag e Switch Mainframe  
Femto Leakage Switch Modul e  
14ch Low Leakag e Switch Mainfram e  
14ch Low Leakag e Switch Modul e  
Agilent B2200A  
Agilent B2210A  
Agilent B2201A  
Agilent B2211A  
This declaration covers all options of the above product(s )  
complies with the essential requirements of the following applicable European Directives, and carries  
the CE marking accordingl y:  
Low Voltage Directive (73/23/EEC, amended by 93/68/EEC )  
EMC Direc tive (89/336/EEC, amended by 93/68/EEC)  
and conforms with the following product standard s  
EMC  
Standar d  
Limit  
IEC 61326:2002/ EN 61326:02  
CISPR 11:1999+A2 / EN 55011:1998+A1  
IEC 61000-4-2:2001 / EN 61000-4-2:1995+A1+A2  
IEC 61000-4-3:2002 / EN 61000-4-3:1996+A1+A2  
IEC 61000-4-4:1995+A1+A2 / EN 61000-4-4:1995+A1  
IEC 61000-4-5:2001 / EN 61000-4-5:1995+A1  
IEC 61000-4-6:2001 / EN 61000-4-6:1996+A1  
IEC 61000-4-11:2001 / EN 61000-4-11:1994+A1  
Group1 Class A  
4 kV CD, 8 kV AD  
3 V/m, 80-1000 MHz  
0.5 kV signal lines, 1 kV power lines  
0.5 kV line-line, 1 kV line-ground  
3 V, 0.15-80 MHz  
1cycle, 100%  
Canada: ICES-001:1998  
Australia/New Zealand: AS/NZS 2064.1  
The product was tested in a typical configuration with Agilent Technologies test systems.  
IEC 61010-1:2001 / EN 61010-1:2001  
Safety  
Canada: CSA C22.2 No. 1010.1:1992, NRTL/C  
Supplementary Information:  
This DoC applies to above-listed products placed on the EU market after:  
Sep 24, 2004  
Date  
Toshiyuki Kawaj i  
QA Manager  
Agilent Technologie s  
Herstellerbescheinigung  
GEÄUSCHEMISSION  
Lpa < 70 dB  
am Arbeitsplatz  
normaler Betrieb  
nach DIN 45635 T. 19  
Manufacturer’s Declaration  
ACOUSTIC NOISE EMISSION  
Lpa < 70 dB  
operator position  
normal operation  
per ISO 7779  
NOTE  
This ISM device complies with Canadian ICES-001.  
Cet appareil ISM est conforme ?Hla norme NMB-001 du Canada.  
This product complies with the WEEE Directive (2002/96/EC) marking  
requirements. The affixed label indicates that you must not discard this electrical/  
electronic product in domestic household waste.  
Product Category: With reference to the equipment types in the WEEE Directive  
Annex I, this product is classed as a “Monitoring and Control instrumentation”  
product.  
Do not dispose in domestic household waste.  
To return unwanted products, contact your local Agilent office, or see  
www.agilent.com/environment/product/ for more information.  
Microsoft, Windows, Windows NT, Visual C++, and Visual Basic are registered trademarks of Microsoft Corporation.  
Borland C++ Builder is registered trademark of International, Inc. LabWindows and LabVIEW are registered  
trademarks of National Instruments Corporation. All other trademarks are the property of their respective owners.  
Safety Summary  
The following general safety precautions must be observed during all phases of  
operation, service, and repair of this instrument. Failure to comply with these  
precautions or with specific warnings elsewhere in this manual may impair the  
protections provided by the equipment. In addition, it violates safety standards of  
design, manufacture, and intended use of the instrument. Agilent Technologies, Inc.  
assumes no liability for customer’s failure to comply with these requirements.  
NOTE  
Agilent B2200 complies with INSTALLATION CATEGORY II for mains input and  
INSTALLATION CATEGORY I for measurement input terminals, and  
POLLUTION DEGREE 2 defined in IEC 1010-1.  
Agilent B2200 is INDOOR USE products.  
GROUND THE INSTRUMENT  
This is Safety Class I instrument. To minimize shock hazard, the instrument  
chassis and cabinet must be connected to an electrical ground. The power  
terminal and the power cable must meet International Electrotechnical  
Commission (IEC) safety standards.  
DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE  
Do not operate the instrument in the presence of flammable gases or fumes.  
Operation of any electrical instrument in such an environment constitutes a  
definite safety hazard.  
KEEP AWAY FROM LIVE CIRCUITS  
Operation personnel must not remove instrument covers. Component  
replacement and internal adjustments must be made by qualified maintenance  
personnel. Do not replace components with power cable connected. Under  
certain conditions, dangerous voltages may exist even with the power cable  
removed. To avoid injuries, always disconnect power and discharge circuits  
before touching them.  
DO NOT SERVICE OR ADJUST ALONE  
Do not attempt internal service or adjustment unless another person, capable of  
rendering first aid and resuscitation, is present.  
DO NOT SUBSTITUTE PARTS OR MODIFY INSTRUMENT  
Because of the danger of introducing additional hazards, do not install substitute  
parts or perform any unauthorized modification to the instrument. Return the  
instrument to a Agilent Technologies Sales and Service Office for services and  
repair to ensure that safety features are maintained.  
DANGEROUS PROCEDURE WARNINGS  
Warnings, such as example below, precede potentially dangerous procedures  
throughout this manual. Instructions contained in the warnings must be  
followed.  
WARNING  
Dangerous Voltage, capable of causing death, are present in this instrument.  
Use extreme caution when handling, testing, and adjusting.  
Safety Symbols  
The general definitions of safety symbols used on equipment or in manuals are  
listed below.  
Instruction manual symbol: the product will be marked with this symbol when it is  
necessary for the user to refer to the instruction manual in order to protect against  
damage to the instrument.  
Indicates dangerous voltage and potential for electrical shock. Do not touch  
terminals that have this symbol when insrument is on.  
Affixed to product containing static sensitive devices--use anti-static handling  
procedures to prevent electrostatic discharge damage to component.  
Protective conductor terminal. For protection against electrical shock in case of a  
fault. Used with field wiring terminals to indicate the terminal which must be  
connected to ground before operating equipment.  
Frame or chassis terminal. A connection to the frame (chassis) of the equipment  
which normally includes all exposed metal structures.  
Indicates earth (ground) terminal.  
Alternating current.  
Direct current.  
ON (Supply).  
OFF (Supply).  
STANDBY (Supply).  
Means INSTALLATION CATEGORY I. Measurement terminals on the rear panel  
comply with INSTALLATION CATEGORY I.  
CAT 1  
WARNING  
CAUTION  
The warning sign denotes a hazard. It calls attention to a procedure, practice,  
condition or the like, which, if not correctly performed or adhered to, could result in  
injury or death to personal.  
The caution sign denotes a hazard. It calls attention to an operating procedure,  
practice, condition or the like, which, if not correctly performed or adhered to, could  
result in damage to or destruction of part or all of the product.  
In This Manual  
This manual is a user’s guide for Agilent B2200A/B2201A Switch Mainframe, and  
consists of the following chapters:  
Introduction  
Describes an overview and specifications of the Agilent B2200 series.  
Installation  
Describes how to install the Agilent B2200 and how to setup the input/output.  
Front Panel Operation  
Explains the front panel operation and the switch control functions of the  
Agilent B2200, also provides the reference information of the front panel keys  
and display.  
Programming  
Explains the programming examples that control the Agilent B2200.  
SCPI Command Reference  
Describes the all commands of the Agilent B2200 SCPI commands.  
VXIplug&play Driver  
Describes the all functions of the Agilent B2200 VXIplug&play driver.  
Error Messages  
Lists and describes the error messages for the Agilent B2200.  
Text Conventions  
The following text conventions are used in this manual:  
Screen Text  
Represents text that appears on screen of the controller.  
Refers to a related document, or is used for emphasis.  
Italic  
Contents  
1. Introduction  
Agilent B2200 Series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3  
Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4  
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4  
Rear Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6  
General Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-11  
Supplemental Information for B2200A/B2210A. . . . . . . . . . . . . . . . . . . . . . . 1-13  
Supplemental Information for B2201A/B2211A . . . . . . . . . . . . . . . . . . . . . . 1-14  
Power Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3  
Power Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3  
Operating Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5  
Storage and Shipping Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5  
To Set the GPIB Address . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8  
To Connect the GPIB Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8  
To Install the Switch Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9  
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9  
To Install the Blank Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9  
Agilent B2200 User’s Guide, Edition 2  
Contents  
Connector Plates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12  
To Make Connections to DUT Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13  
Input Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-20  
To Use Couple Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9  
To Display Firmware Revision . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9  
To Display Module Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9  
To Read Error Message . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10  
To Set Beeper. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10  
Agilent B2200 User’s Guide, Edition 2  
Contents  
Ground Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-18  
Couple Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-20  
LED Matrix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-22  
LCD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-23  
Function key group. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-26  
4. Programming  
Switch Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5  
Programming Examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7  
Connecting Input-Output Paths . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8  
Using Bias Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10  
Using Ground Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12  
Agilent B2200 User’s Guide, Edition 2  
Contents  
Capacitance Compensation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-20  
:DIAGnostic:TEST:CARD:STATe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-22  
:DIAGnostic:TEST:FRAMe:CLEar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-22  
:DIAGnostic:TEST:FRAMe[:EXECute]? . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-23  
:DIAGnostic:TEST:FRAMe:STATe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-24  
[:ROUTe]:AGND:CHANnel:DISable:CARD . . . . . . . . . . . . . . . . . . . . . . . . . 5-24  
Agilent B2200 User’s Guide, Edition 2  
Contents  
:SYSTem:CPON . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-46  
:SYSTem:CTYPe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-47  
:SYSTem:DISPlay:LCD. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-47  
:SYSTem:DISPlay:LED. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-48  
:SYSTem:DISPlay:STRing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-48  
Agilent B2200 User’s Guide, Edition 2  
Contents  
Status Byte Register . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-55  
Standard Event Status Enable Register . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-59  
Output Queue . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-60  
System Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3  
agb220xa_close . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-10  
agb220xa_closeCard_Q . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-10  
agb220xa_closeList . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-11  
agb220xa_closeList_Q. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-12  
agb220xa_cmd . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-12  
Agilent B2200 User’s Guide, Edition 2  
Contents  
agb220xa_init . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-27  
agb220xa_opc_Q . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-27  
agb220xa_openCard. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-28  
agb220xa_openList. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-28  
agb220xa_openList_Q . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-29  
Agilent B2200 User’s Guide, Edition 2  
Contents  
agb220xa_testClear . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-31  
Device-Dependent Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-8  
Query Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-9  
B2200 Specific Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-10  
B2200 Channel Related Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-11  
B2200 Card/Mode/Port Related Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-12  
Agilent B2200 User’s Guide, Edition 2  
1
Introduction  
 
Introduction  
This chapter describes the basic functions and features of the Agilent  
B2200A/B2201A (Agilent B2200 series), and consists of the following sections:  
“Rear Panel”  
“Switch Modules”  
“Specifications”  
“Accessories and Options”  
1-2  
Agilent B2200 User’s Guide, Edition 2  
Introduction  
Agilent B2200 Series  
Agilent B2200 Series  
Agilent B2200 series is a switching matrix designed for semiconductor dc  
characteristics measurement applications. The Agilent B2200 series has 14 input  
ports and four card slots for the switch modules (plug-in cards), and can configure a  
12, 24, 36, or 48 outputs switching matrix. The Agilent B2200A/B2201A supports  
the following dedicated switch module.  
Agilent B2210A fA Leakage Switch Module  
Dedicated for the Agilent B2200A mainframe.  
Agilent B2211A 14ch Low Leakage Switch Module  
Dedicated for the Agilent B2201A mainframe.  
NOTE  
Mixed configuration of the switch modules is not supported.  
Basic functions of the Agilent B2200 series are listed below.  
SMU inputs (for source monitor units): 8 ports  
AUX inputs (multipurpose inputs): 6 ports  
Outputs: 12, 24, 36, or 48 ports  
Connection setup and status monitor on the front panel  
Status display on the LED matrix display  
Connection setup by the Light pen  
Automatic control through GPIB interface  
Auto ground function  
Self-test, relay function test  
Relay cleaning  
Agilent B2200 User’s Guide, Edition 2  
1-3  
 
Introduction  
Front Panel  
Front Panel  
The Agilent B2200 series provides the front panel keys, the LCD, and the LED  
matrix display for the status monitor and connection setup.  
Line switch  
Used to turn the Agilent B2200 on or off.  
LCD  
Used to monitor the status and set the connection. See Chapter 3 for the details.  
Front panel keys  
Used to set the Agilent B2200. See Chapter 3 for the details.  
LED matrix display  
Displays the status of the matrix switches. Also used to set the matrix connection  
with the light pen.  
SMU Input  
Inputs for the source monitor unit (SMU). Eight input ports. Up to four kelvin  
inputs are available. Triaxial BNC connector.  
AUX Input  
Multipurpose inputs. Six input ports. Coaxial BNC connector. CMH and CML  
terminals are the input ports for the capacitance measurement.  
The input port 12 (AUX Input 12) is the default ground input port. If you assign  
the port as the ground input port, this ground port will be internally connected to  
the ground when the ground mode is set to ON. In this case, open this connector.  
Light Pen  
Connector for the Agilent 16443A Light pen. Used to set the matrix connection  
with the LED matrix display.  
CAUTION  
For the SMU Input terminals, the maximum measurement voltage/current/voltage  
between terminals are ±200 V/1 A/300 V, respectively. And for the AUX Input  
terminals, they are ±100 V/0.5 A/100 V.  
Do not apply an input signal over these limits to the input terminals. If you do, the  
Agilent B2200 will be damaged. If you use a bias source that has current limit  
capability, set the bias source current limit.  
1-4  
Agilent B2200 User’s Guide, Edition 2  
   
Introduction  
Front Panel  
Figure 1-1  
Front Panel View  
Agilent B2200 User’s Guide, Edition 2  
1-5  
Introduction  
Rear Panel  
Rear Panel  
The Agilent B2200 series has four card slots for the switch modules, the GPIB  
interface and so on.  
Card slots  
For the Agilent B2200A mainframe, the Agilent B2210A cards are installed.  
For the Agilent B2201A mainframe, the Agilent B2211A cards are installed.  
Mixed configuration of the switch modules is not supported.  
GPIB interface  
Use an Agilent 82357A USB/GPIB interface or Agilent 10833A/B/C/D GPIB  
cable to connect to an external computer or equipment.  
Serial number  
You need this serial number when using the Agilent Technologies telephone  
assistance program.  
LINE input receptacle  
AC power cable is connected to this receptacle.  
1-6  
Agilent B2200 User’s Guide, Edition 2  
 
Introduction  
Rear Panel  
Figure 1-2  
Rear Panel View  
Agilent B2200 User’s Guide, Edition 2  
1-7  
Introduction  
Switch Modules  
Switch Modules  
The Agilent B2200A and B2201A support the dedicated switch module, Agilent  
B2210A and B2211A, respectively.  
By installing the modules, the module inputs will be connected internally to the  
front panel input connectors. And 12 output connectors will face the rear panel. The  
type of the output connectors is the triaxial BNC. Up to six kelvin outputs are  
available.  
Image of the output terminals and the block diagram of the switch module are  
shown in Figure 1-3 and Figure 1-4.  
Figure 1-3  
Switch Module Output Terminals  
WARNING  
NOTE  
Do not touch the force and guard terminals of the output connectors while the  
Agilent B2200 is turned on. Dangerous voltages up to the maximum input  
voltage may be present at the output connectors.  
Mixed configuration of the switch modules is not supported.  
1-8  
Agilent B2200 User’s Guide, Edition 2  
     
Introduction  
Switch Modules  
Figure 1-4  
Switch Module Block Diagram  
Output  
1
2
3
4
5
9
6
7
8
10  
11  
12  
Input  
1
2
3
4
5
6
7
8
9
10  
11  
12  
13 CMH  
14 CML  
Agilent B2200 User’s Guide, Edition 2  
1-9  
 
Introduction  
Specifications  
Specifications  
This section lists specifications and supplemental information for the Agilent B2200  
series.  
“General Specifications”  
“Switch Modules”  
“Supplemental Information for B2200A/B2210A”  
“Supplemental Information for B2201A/B2211A”  
The specifications are the performance standards or limits against which these units  
have been tested. The supplemental information is not warranted, but provides  
useful information about functions and performance.  
If not noted otherwise, the conditions for specifications and supplemental  
information are as follows:  
Temperature: 23 °C ± 5 °C  
Humidity: 5 % to 60 % R.H.  
1-10  
Agilent B2200 User’s Guide, Edition 2  
 
Introduction  
Specifications  
General Specifications  
Temperature  
range:  
Operating: 5 °C to 35 °C  
Storage: 20 °C to 70 °C  
Humidity range:  
Operating: 5 % to 70 % R.H., non-condensing  
Storage:  
< 80 % R.H. at 35 °C, < 60 % R.H. at 65 °C, non-condensing (B2200A/B2210A)  
< 80 % R.H. at 65 °C, non-condensing (B2201A/B2211A)  
Altitude:  
Operating: 0 to 2,000 m (6,500 ft)  
Storage: 0 to 15,240 m (50,000 ft)  
Regulatory  
compliance:  
Safety: CSA C22.2 No. 1010.1/IEC 1010-1  
EMC: CISPR 11 Group 1 class A&EN50082-1  
Power  
90 to 264 V (continuous), 47 to 63 Hz, 2 A/200 VA maximum  
requirement:  
Number of slots:  
Dimensions:  
4 slots for 48 mm height switch module  
B2200A/B2201A: 430 mm W × 320 mm H × 600 mm D  
B2210A/B2211A: 395 mm W × 48 mm H × 500 mm D  
Weight:  
B2200A/B2201A: approximately 14.0 kg  
B2210A: approximately 5.0 kg  
B2211A: approximately 3.5 kg  
Number of ports:  
SMU input: 8 triaxial ports (with Guard)  
AUX input: 6 coaxial ports (two ports for capacitance measurement)  
Output: 12 triaxial ports (with Guard), maximum 48 ports  
Agilent B2200 User’s Guide, Edition 2  
1-11  
 
Introduction  
Specifications  
Switch Modules  
Table 1-1 lists the specifications when the switch module is installed in the  
mainframe. In the table, the conditions are as follows:  
Conditions: 23 °C ± 5 °C, 5 % to 60 % R.H.  
Table 1-1  
Agilent B2210A/B2211A Switch Module Specifications  
B2210A  
B2211A  
Max. current rating (A)  
SMU input  
1.0  
0.5  
1.0  
0.5  
AUX input  
Max. voltage rating (V)  
SMU input, channel to common  
SMU input, channel to channel  
AUX input, channel to common  
AUX input, channel to channel  
Closed channel residual resistance ()  
SMU input  
200  
300  
100  
100  
200  
300  
100  
100  
0.6  
1.5  
0.6  
1.5  
AUX input  
Channel isolation resistance ()  
SMU input  
1014  
109  
5 × 1013  
109  
AUX input  
The Agilent B2200A mainframe supports the Agilent B2210A module.  
The Agilent B2201A mainframe supports the Agilent B2211A module.  
Mixed configuration of the switch modules is not supported.  
1-12  
Agilent B2200 User’s Guide, Edition 2  
   
Introduction  
Specifications  
Supplemental Information for B2200A/B2210A  
Offset current 1  
< 10 fA (typical 3 fA) (SMU input)  
IM noise (RMS) 2  
0.6 fA (SMU input)  
Channel crosstalk capacitance < 1 pF/channel (SMU input)  
< 3 pF/channel (AUX input)  
Offset voltage  
< 50 µV (SMU input)  
< 80 µV (AUX input)  
2.0 seconds  
Settling time 3  
Bandwidth (at -3dB)  
Guard capacitance 4  
30 MHz (SMU input)  
< 145 pF (SMU input)  
Additional C measurement error 5 < ± 1 % + 0.2 pF (SMU input)  
1. When the voltage applied to all input-output channels is 0 V.  
2. Measured by the Agilent 4156C with the integration time setting 100 PLC.  
When the voltage applied to all other input-output paths is 0 V.  
3. The time until the measurement value settles to within 50 fA of the final  
value. When the applied voltage is 10 V.  
4. When four modules have been installed in the mainframe. And when only  
one input-output path is made.  
5. For the measurement less than 1000 pF at the frequency 1 kHz to 1 MHz,  
using the Agilent 4284A with 3 m cable. After the compensation using the  
capacitance compensation algorithm (a function of VXIplug&play driver).  
Agilent B2200 User’s Guide, Edition 2  
1-13  
 
Introduction  
Specifications  
Supplemental Information for B2201A/B2211A  
Offset current 1  
< 50 fA (SMU input)  
IM noise (RMS) 2  
5 fA (SMU input)  
Channel crosstalk capacitance < 0.5 pF/channel (SMU input)  
< 3 pF/channel (AUX input)  
Offset voltage  
< 80 µV (SMU input)  
< 100 µV (AUX input)  
2.0 seconds  
Settling time 3  
Bandwidth (at -3dB)  
Guard capacitance 4  
30 MHz (SMU input)  
< 145 pF (SMU input)  
Additional C measurement error 5 < ± 1 % + 0.2 pF (SMU input)  
1. When the voltage applied to all input-output paths is 0 V.  
2. Measured by the Agilent 4156C with the integration time setting 100 PLC.  
When the voltage applied to all other input-output paths is 0 V.  
3. The time until the measurement value settles to within 300 fA of the final  
value. When the applied voltage is 10 V.  
4. When four modules have been installed in the mainframe. And when only  
one input-output path is made.  
5. For the measurement less than 1000 pF at the frequency 1 kHz to 1 MHz,  
using the Agilent 4284A with 3 m cable. After the compensation using the  
capacitance compensation algorithm (a function of VXIplug&play driver).  
1-14  
Agilent B2200 User’s Guide, Edition 2  
 
Introduction  
Accessories and Options  
Accessories and Options  
Agilent B2200 is furnished with the following accessories.  
Power cable, 1 ea.  
Operation summary sheet, 1 ea.  
Manual CD-ROM, 1 ea.  
Software CD-ROM, 1 ea.  
Stores the Agilent B2200 VXIplug&play driver.  
The Agilent B2200 VXIplug&play driver supports Windows XP Professional,  
Windows 2000, Windows NT 4.0, Windows 98, and Windows 95.  
Moisture-proof and dehumidifying packing kit, 1 set  
Table 1-2 lists the options and accessories available for the Agilent B2200.  
Agilent B2200 User’s Guide, Edition 2  
1-15  
 
Introduction  
Accessories and Options  
Table 1-2  
Options and Accessories  
Option Item  
Model  
Number  
Description  
B2200A  
fA Leakage Switch Mainframe  
B2200A-UK6  
Commercial cal. certificate w/ test data  
B2200A-ABA Manual set, English  
B2200A-ABJ  
Manual set, Japanese  
B2201A  
14ch Low Leakage Switch Mainframe  
Commercial cal. certificate w/ test data  
B2201A-UK6  
B2201A-ABA Manual set, English  
B2201A-ABJ  
Manual set, Japanese  
B2210A  
B2211A  
16443A  
16493H  
fA Leakage Switch Module (for B2200A)  
14ch Low Leakage Switch Module (for B2201A)  
Light pen  
GNDU cable (between 41501/4142 and 16495F/G)  
1.5 m length  
16493H-001  
16493H-002  
3 m length  
16493J  
Interlock cable  
(between E5260/E5270/4155/4156 and 16495F/G)  
16493J-001  
16493J-002  
16493J-003  
1.5 m length  
3 m length  
5 m length  
16493K  
16493L  
Kelvin triaxial cable  
(between B2200 inputs and E5260/E5270/4156/41501)  
16493K-001  
16493K-002  
1.5 m length  
3 m length  
GNDU cable (between E5260/E5270 and 16495F/G)  
16493L-001  
16493L-002  
16493L-003  
1.5 m length  
3 m length  
5 m length  
16493N  
GNDU cable for Kelvin connection  
(between B2200 inputs and E5260/E5270/41501/4142)  
1-16  
Agilent B2200 User’s Guide, Edition 2  
 
Introduction  
Accessories and Options  
Model  
Number  
Option Item  
Description  
16494A  
Triaxial cable  
1.5 m length  
3 m length  
16494A-001  
16494A-002  
16494A-003  
16494A-005  
80 cm length  
4 m length  
16494B  
Kelvin triaxial cable  
(between B2200 inputs and 4142B, between B2210/B2211  
outputs and 16495F/G)  
16494B-001  
16494B-002  
16494B-003  
1.5 m length  
3 m length  
80 cm length  
16494C  
16494F  
Kelvin triaxial cable  
(between B2210/B2211 outputs and B2220A)  
16494C-001  
16494C-002  
16494C-005  
1.5 m length  
3 m length  
4 m length  
CMU Input cable, 2 m  
(between B2200 CMH/CML and 4-terminal pairs connectors)  
16495E  
16495F  
Half size blank plate  
Half size connector plate with 12×triaxial, intlk, GNDU  
Bulkhead feedthrough connectors (female to female)  
Connectors to contacts for soldering  
16495F-001  
16495F-002  
16495G  
Full size connector plate with 24×triaxial, intlk, GNDU  
Bulkhead feedthrough connectors (female to female)  
Connectors to contacts for soldering  
16495G-001  
16495G-002  
Agilent B2200 User’s Guide, Edition 2  
1-17  
Introduction  
Accessories and Options  
1-18  
Agilent B2200 User’s Guide, Edition 2  
2
Installation  
 
Installation  
This chapter describes requirements to install Agilent B2200 and the tasks for  
installation, and is organized into the following three sections:  
“Requirements”  
“Inspection”  
“Self-Test”  
“Output Connections”  
“Input Connections”  
“Measurement Cable Length”  
“Maintenance”  
WARNING  
The maximum input voltage of the Agilent B2200 is ±200 V. And dangerous  
voltages may be present at the output terminals. To prevent electric shock, you  
must observe the following safety precautions when using the Agilent B2200.  
• Use a three-conductor ac power cable to connect cabinet (if used) and the  
Agilent B2200 to an electric ground (safety ground).  
• If you need to touch the force and guard terminals of the output connector,  
turn off the Agilent B2200 and discharge any capacitors whenever possible.  
• Warn workers around the B2200 about dangerous conditions.  
2-2  
Agilent B2200 User’s Guide, Edition 2  
 
Installation  
Requirements  
Requirements  
This section describes the following requirements for the Agilent B2200.  
“Power Requirements”  
“Power Cable”  
“Operating Environment”  
“Storage and Shipping Environment”  
Power Requirements  
CAUTION  
Before applying ac line power to the Agilent B2200, ensure that the correct power  
cable is used.  
The Agilent B2200 can operate from any single-phase ac power source supplying 90  
to 264 V in the frequency range from 47 to 63 Hz. The maximum power  
consumption is 2 A/200 VA.  
Power Cable  
In accordance with international safety standards, this instrument is equipped with a  
three-wire power cable. When connected to an appropriate ac power outlet, this  
cable grounds the instrument frame. The type of power cable shipped with each  
instrument depends on the country of destination. See Table 2-1 for the part numbers  
of the power cables available.  
If the plug on the cable does not fit the power outlet, or the cable is to be attached to  
a terminal block, cut the cable at the plug end and re-wire it. This work should be  
performed by a qualified electrician all local electrical codes being strictly observed.  
The color coding used in the cable will depend on the cable supplied. If a new plug  
is to be connected, it must meet local safety requirements and include the following  
features:  
Adequate load-carrying capacity.  
Ground connection.  
Cable clamp.  
Agilent B2200 User’s Guide, Edition 2  
2-3  
     
Installation  
Requirements  
Table 2-1  
Power Cable  
Plug: CEE 7 Standard  
Sheet VII, 250 V, 10 A  
Cable: 8120-1689  
Plug: AS 3112, 250 V,  
10 A  
Cable: 8120-1369  
Plug: BS 1363/A,  
250 V, 10 A  
Cable: 8120-1351  
Plug: SEV Type 12,  
250 V, 10 A  
Cable: 8120-2104  
Plug: NEMA 5-15P,  
125 V, 10 A  
Cable: 8120-1378  
Plug: NEMA 6-15P,  
250 V, 6 A  
Cable: 8120-0698  
Plug: JIS C 8303,  
125 V, 12 A  
Cable: 8120-4753  
Plug: SR 107-2-D,  
250 V, 10 A  
Cable: 8120-2956  
Plug: IEC 83-B1,  
250 V, 10 A  
Cable: 8120-4211  
Plug: CEI 23-16, 250 V,  
10 A  
Cable: 8120-6978  
Plug: Argentine  
Resolution 63, Annex  
IV, 250 V, 10 A  
Plug: GB 1002, 250 V,  
10 A  
Cable: 8120-8376  
Cable: 8120-6870  
2-4  
Agilent B2200 User’s Guide, Edition 2  
 
Installation  
Requirements  
WARNING  
For protection from electrical shock, the power cable ground must not be  
defeated.  
Operating Environment  
The Agilent B2200 must be operated within the following environmental  
conditions:  
Temperature: 5 °C to 35 °C  
Humidity: 5 % to 70 % R.H., non-condensing  
Altitude: 0 m to 2,000 m  
Storage and Shipping Environment  
The Agilent B2200 should be stored or shipped in environments within the  
following limits:  
Temperature: -20 °C to 70 °C  
Humidity:  
B2200A/B2210A:  
< 80 % R.H. (at 35 °C), < 60 % R.H. (at 65 °C), non-condensing  
B2201A/B2211A:  
< 80 % R.H. (at 65 °C), non-condensing  
Altitude: 0 m to 15,240 m  
CAUTION  
When storing the Agilent B2200  
Protect the Agilent B2200 from temperature extremes to prevent condensation from  
forming inside the Agilent B2200. If condensation occurs, the Agilent B2200 may  
damage or may not satisfy the specifications. If the Agilent B2200 was placed at the  
high temperature and high humidity environment long time, the Agilent B2200 may  
degrade its performance.  
NOTE  
When shipping the Agilent B2200  
The Agilent B2200 must be packed in the certain packing materials for protection  
from damage when it is shipped. Cover the Agilent B2200 by using the  
moisture-proof and dehumidifying packing material (furnished), and pack it by  
using packing materials such as carton box and cushioning materials.  
Agilent B2200 User’s Guide, Edition 2  
2-5  
   
Installation  
Inspection  
Inspection  
CAUTION  
Before Opening Packing Materials  
The Agilent B2200 contains the condensation sensitive electronic parts. The  
condensation will have a negative impact on the Agilent B2200 to operate normally.  
Do not open the packing materials, and leave the Agilent B2200 to acclimate it to  
the installation environment (temperature and humidity). If it is opened without  
enough acclimation, the Agilent B2200 may damage.  
When you open the box that contains the Agilent B2200, check the following:  
1. Before unpacking any components, inspect all boxes for any signs of damage  
that might have occurred during shipment such as:  
Dents  
Scratches  
Cuts  
Water marks  
2. When you open the boxes that contain the Agilent B2200, check the components  
against the contents lists that are attached to the boxes.  
If anything is wrong, notify your local Agilent Technologies sales office.  
2-6  
Agilent B2200 User’s Guide, Edition 2  
 
Installation  
Installing the B2200  
Installing the B2200  
This section describes how to install the Agilent B2200.  
1. See “Requirements” on page 2-3, and determine the installation location.  
2. Unpack the Agilent B2200 and place it at the installation site.  
3. Plug the power cable into the Line input receptacle at the rear panel.  
4. Plug the power cable into the power receptacle.  
5. Perform the self-test. See “Self-Test” on page 2-10.  
6. If you control the Agilent B2200 by using an external computer:  
Set the GPIB address. See “To Set the GPIB Address” on page 2-8 and “To  
Connect the GPIB Cable” on page 2-8.  
7. If you have to add or replace the switch module:  
Turn the Agilent B2200 off, remove the power cable, and install the switch  
module. See “To Install the Switch Module” on page 2-9 and “To Install the  
Blank Panel” on page 2-9.  
NOTE  
Plug-in module is shipped from the factory after the specifications are confirmed.  
Agilent can guarantee that the modules will function and that the module  
performance is designed to meet its specifications. However, if you have any  
trouble, contact your nearest Agilent Technologies service center.  
If you require that the modules be guaranteed to their specifications (for ISO  
compliance, etc.), then you must perform a calibration on the instrument (modules  
and mainframe together). For calibration, contact your nearest Agilent Technologies  
service center.  
Agilent B2200 User’s Guide, Edition 2  
2-7  
 
Installation  
Installing the B2200  
To Set the GPIB Address  
Every device on the GPIB bus must have a unique address. If you need to change  
the GPIB address, turn the Agilent B2200 on and perform the following procedure.  
The new GPIB address is recognized only at power on. The Agilent B2200 leaves  
the factory with the GPIB address set to 22.  
1. Press the Menu key.  
2. Move the cursor to CONFIG, then press the Enter key.  
3. Move the cursor to ADDRESS, then press the Enter key. The following message  
will appear on the LCD. XX will be 0 to 30.  
GPIB Address = XX  
4. Press the arrow key (up or down) to specify the desired GPIB address, then press  
the Enter key.  
5. Press the Enter key twice.  
6. Turn the Agilent B2200 off, and then turn it on again.  
To Connect the GPIB Cable  
To connect the instrument with a computer or peripheral device via GPIB (IEEE  
Std. 488), connect an GPIB cable between the GPIB connector on the instrument  
rear panel and the GPIB connector on the peripheral device.  
Including the controller, a total of 15 GPIB interfaces can be connected on the same  
GPIB bus. The following are rules for connecting GPIB interfaces:  
If total number of interfaces 10:  
Max. total cable length = number of interfaces × 2 m.  
If total number of interfaces > 10:  
Max. total cable length = 20 m.  
Maximum cable length between interfaces is 4 m.  
Star connection and cascade connection are allowed. Loop connection is not  
allowed.  
2-8  
Agilent B2200 User’s Guide, Edition 2  
   
Installation  
Installing the B2200  
To Install the Switch Module  
WARNING  
CAUTION  
To prevent electrical shock, turn off the mainframe and remove the power  
cable before starting the instruction.  
Be careful about the module pins used for internal connection to the Agilent B2200.  
The pins can be damaged easily.  
Use clean handling and anti-static procedures when removing, configuring, and  
installing the switch modules. The modules contain components that can be  
damaged by static electricity.  
The following procedure explains the module installation and removal.  
1. Turn off the Agilent B2200, then wait at least 10 seconds before you remove or  
install a module.  
2. Loosen the screws on both the left and right edges of a blank panel or a module  
attached to the slot you want to install a new module.  
3. Remove the blank panel or the module.  
4. Align the new module with the left and right slot guide rails. Then the  
component side should be facing up.  
5. Push the module into the slot until you feel it seat firmly into the connector at the  
back of the slot.  
6. Screw in the screws on the left and right edges of the module.  
7. Execute the self-test and the relay test. See “Self-Test” on page 2-10.  
To Install the Blank Panel  
CAUTION  
To prevent thermal damage to the module, be sure that blank panels are installed in  
all unused slots.  
If the blank panel is not installed to cover an unused slot, install the blank panel as  
follows:  
1. Align the blank panel over the unused slot.  
2. Screw in the screws on the left and right edges of the blank panel.  
Agilent B2200 User’s Guide, Edition 2  
2-9  
     
Installation  
Self-Test  
Self-Test  
NOTE  
To confirm the specifications  
The self-test and diagnostics checks the operation of the mainframe and the  
modules. However they cannot confirm if the Agilent B2200 satisfies its  
specifications.  
For verifying the specifications, contact your nearest Agilent Technologies Service  
Center. Trained service personnel will perform calibration (performance  
verification).  
It is recommended to perform calibration once a year at least.  
The following procedure performs the self-test and diagnostics.  
1. Turn on the Agilent B2200.  
The controller test will be performed. If the Agilent B2200 fails the test, contact  
your nearest Agilent Technologies Service Center.  
2. Press the Shift key and the Menu key in this order. Self-test menu will appear on  
the LCD.  
SELF_TEST  
RELAY_TEST  
KEY  
Controller test  
Relay test  
Front panel interface test  
Beeper test  
BEEPER  
LED  
LED matrix test  
Light pen test  
GPIB test  
PEN  
GPIB  
3. Move the cursor to the test item to perform, then press Enter key.  
4. To start the test, move the cursor to EXECUTE, then press Enter key.  
To display the previous test result, move the cursor to RESULT, then press Enter  
key.  
For the details of each test, see “Selftest Menu” on page 3-34.  
2-10  
Agilent B2200 User’s Guide, Edition 2  
 
Installation  
Output Connections  
This section describes how to connect the Agilent B2200 outputs to prober,  
“Connector Plates”  
“To Make Connections to DUT Interface”  
“To Make Interlock Circuit”  
“To Mount Connectors”  
NOTE  
Output Connections  
If you do not use the connector plate for the connection between the output and the  
DUT interface, see “To Mount Connectors” on page 2-18.  
WARNING  
Turn off the Agilent B2200 and all instruments connected. And do not turn  
them on until the connection described in this section is completed. If you  
ignore this warning, you may be exposed to dangerous voltage.  
Output Connectors  
The Agilent B2200 output connectors are the triaxial BNC connector. The input  
signals appears at the output as shown in Figure 2-1 when input-output switching  
path is made. Figure 2-1 shows signals appear on the output connectors with  
non-Kelvin connection. If you make Kelvin connection, the signal of the even  
output connectors is Sense, not Force.  
Figure 2-1  
Output Connector and Output Signal  
Signal (force)  
Signal (force)  
GND  
GUARD  
GND  
(b) When AUX is connected  
(a) When SMU is connected  
Agilent B2200 User’s Guide, Edition 2  
2-11  
     
Installation  
Output Connections  
Connector Plates  
Connector plates (Table 2-2) are used for the connection between the Agilent B2200  
outputs and the DUT interface (prober and so on). To connect to the connector plate,  
use the cable shown in Table 2-3.  
Table 2-2  
Connector Plate  
Agilent Model No.  
Description  
16495F  
Half size connector plate  
16495F-001 has 12 triaxial through connectors (female to female), an Intlk  
connector, and a GNDU connector (triaxial through, female to female). The back  
of the Intlk connector is designed for soldering.  
16495F-002 has 12 triaxial connectors, an Intlk connector, and a GNDU  
connector. The back of each connector is designed for soldering.  
16495G  
16495E  
Full size connector plate  
16495G-001 has 24 triaxial through connectors (female to female), an Intlk  
connector, and a GNDU connector (triaxial through, female to female). The back  
of the Intlk connector is designed for soldering.  
16495G-002 has 24 triaxial connectors, an Intlk connector, and a GNDU  
connector. The back of each connector is designed for soldering.  
Blank plate  
This plate is used to cover openings when you made too big openings for  
mounting the connector plate. You will use this plate to cover openings if you  
mount the half size connector plate in openings made for the full size connector  
plate.  
Table 2-3  
Output Cable  
Agilent Model No.  
Description  
16494A  
16494B  
Triaxial cable (for non-Kelvin connection)  
Kelvin Triaxial cable (for Kelvin connection)  
NOTE  
Installing Connector Plate  
To install the connector plate, refer to Agilent 16495 Installation Guide.  
2-12  
Agilent B2200 User’s Guide, Edition 2  
     
Installation  
Output Connections  
For Kelvin connection, use Kelvin triaxial cable listed in Table 2-3. To make a  
Kelvin output port (1, 3, 5, 7, 9 and 11), couple two E5252A output ports as follows:  
Kelvin Output Port  
Output Port Number  
1
3
1 (Force) and 2 (Sense)  
3 (Force) and 4 (Sense)  
5 (Force) and 6 (Sense)  
7 (Force) and 8 (Sense)  
9 (Force) and 10 (Sense)  
11 (Force) and 12 (Sense)  
5
7
9
11  
To Make Connections to DUT Interface  
This section describes for the connections between the DUT interface and the  
connectors connected to the Agilent B2200 output cables. See Table 2-4.  
NOTE  
Low-Noise Coaxial Cable  
For the extended measurement paths over the connector plate, use low-noise coaxial  
cable (Agilent part number 8121-1191). This cable can maximize the guard effects  
and minimize the impression of the external noise.  
Figure 2-2 shows the cutting example of this cable. Key point is the isolation  
between the conductive layer and the center conductor. So, cut and trim the end of  
the cable as shown in this figure by using a cutter and so on.  
Figure 2-2  
Coaxial Cable Cutting Example  
Cover here using sleeve  
Insulator (black)  
min 2 mm  
Center conductor  
(for Force/Sense signal)  
approx. 10 to 15 mm  
Insulator (clear)  
Conductive layer (black)  
Outer conductor  
(for Guard signal)  
Agilent B2200 User’s Guide, Edition 2  
2-13  
   
Installation  
Output Connections  
Table 2-4  
Connection to the DUT Interface  
Kelvin connections  
non-Kelvin connections  
This connection is available only for the  
Kelvin connectors.  
Following connection is for the Kelvin  
connectors. For the triaxial connectors,  
ignore SENSE terminal, and make  
connection only for FORCE terminal.  
Measurement data will include residual  
resistance from the connection cable.  
This connection can cancel effects of cable  
resistance by connect sense line and force line  
as close as possible to DUT terminal.  
Common  
Common  
Guard  
Guard  
Force  
Force  
to  
DUT  
to  
DUT  
Triaxial  
connector  
Coaxial cable  
Triaxial  
connector  
Coaxial cable  
Common  
Guard  
Common  
Guard  
Sense  
Sense  
Wire  
Insulator  
Plate  
Insulator  
Plate  
CAUTION  
Never connect the guard terminal to any output, including circuit common, chassis  
ground, or the guard terminal of any other unit. Doing so may result in an  
emergency condition.  
For highly accurate current forcing and measurements while minimizing leakage,  
surround all force and sense lines from SMU by a guard as far as possible, and make  
cables stable by taping.  
For reducing capacitance measurement error, tape the cables to any grounded  
materials such as shielding box. If you use probe card, grounding the probe card also  
reduces capacitance measurement error.  
2-14  
Agilent B2200 User’s Guide, Edition 2  
 
Installation  
Output Connections  
To Make Interlock Circuit  
The interlock circuit is to prevent electric shock when touching measurement  
terminals. You must install an interlock circuit on shielding box to prevent  
dangerous voltages when door of the shielding box is open.  
Figure 2-3 shows the pin assignments of the interlock connector.  
Figure 2-3  
Interlock Connector Pin Assignments  
WARNING  
Dangerous voltages of up to the maximum voltage of SMUs may be present at  
force, guard, and sense terminals when the interlock terminals are shorted.  
Install the interlock circuit as follows.  
1. Mount two mechanical switches on your shielding box, so that the switches  
close when the door of the shielding box is closed, and open when the door is  
opened. For the recommended parts and the dimensions of the switch, see Figure  
2-4 and Figure 2-5.  
2. Use wire to connect the two switches in series between pin number 1 and 2 (or  
3) of the interlock connector. See Figure 2-3.  
For example, Agilent 4155/4156 is connected to the interlock circuit, it cannot force  
more than ±40 V when the door is open. When door is closed, it can force more than  
±40 V.  
Agilent B2200 User’s Guide, Edition 2  
2-15  
   
Installation  
Output Connections  
Figure 2-4  
Dimensions of Interlock Switch (Agilent part number 3101-0302)  
6.35  
59.4  
2.8  
8.1  
3.1  
NC  
NO  
COM  
3.1  
22.2  
27.8  
10  
Units: mm  
37.8  
UGI01011,85x60  
Figure 2-5  
Dimensions of Interlock Switch (Agilent part number 3101-3241)  
10.2  
4.3  
27.5  
10.9  
3.1  
3.2  
Switch off  
Switch on  
2.0  
15.9  
3.4  
Units: mm  
UGI01012,85x60  
27.8  
6.8  
To Install LED Circuit  
To install LED circuit on your shielding box, do following:  
1. Mount LED on your shielding box. For the recommended parts and the  
dimensions of the LED, see Figure 2-6.  
2. Use wire to connect the LED between pin 4 and pin 5 (or 6) of the interlock  
connector. See Figure 2-3.  
2-16  
Agilent B2200 User’s Guide, Edition 2  
   
Installation  
Output Connections  
The 4155/4156 semiconductor parameter analyzer's Intlk connector provides the  
interlock signal and a LED drive signal. If a LED is connected between pin 4 and  
pin 5 (or 6) of the interlock connector, the LED lights to indicate high voltage output  
when more than ±40 V is forced from an SMU in the 4155/4156.  
Figure 2-6  
Dimensions of LED (Agilent part number 1450-0641)  
6
5
5.6  
Units: mm  
To Connect Interlock Circuit to Instrument  
Before beginning the measurement, connect the interlock circuit to the instrument’s  
interlock connector as follows.  
For the instruments which has a BNC-Type interlock connector:  
1. Get the following parts.  
Agilent 16493J Interlock cable, 1 ea.  
Agilent 16435A Interlock cable adapter, 1 ea.  
2. Connect the 16493J interlock cable between the interlock circuit and the  
16435A adapter.  
3. Connect the BNC cable (furnished with the adapter) between the adapter and  
the instrument’s interlock connector.  
For Agilent 4155/4156/E5260/E5270:  
Connect the 16493J interlock cable between the interlock circuit and the  
instrument’s interlock connector. You can connect it directly without using any  
adapter.  
Agilent B2200 User’s Guide, Edition 2  
2-17  
 
Installation  
Output Connections  
NOTE  
To Check Interlock Circuit  
If you use the 4155/4156, you can easily check the interlock circuit as follows:  
1. Connect the Intlk connector of the 4155/4156 to your interlock circuit.  
2. Press System front-panel key, then select CALIB/DIAG primary softkey to  
display the SYSTEM: SELF-CALIBRATION/DIAGNOSTICS page.  
3. In the CALIB/DIAG field, select DIAG secondary softkey.  
4. In the CATEGORY field, select I/O PERIPH secondary softkey.  
5. Move pointer to the 403 (INT.) Interlock LEDfield.  
6. Select EXECUTE secondary softkey.  
7. Confirm the following:  
LED turns on within 1 sec from when interlock circuit is shorted.  
LED turns off within 1 sec from when interlock circuit is open.  
To stop the interlock test, select STOP secondary softkey.  
To Mount Connectors  
This section provides the information needed to mount connectors for the Agilent  
B2200 output cable connections on the shielding panel. See this section if you do  
not use the connector plate.  
1. Get the appropriate parts for your situation. Refer to Table 2-5.  
2. Make holes and mount the connectors. Refer to Table 2-6.  
For kelvin connections, use the Kelvin cables. The Kelvin cable requires the  
kelvin triaxial connector which has two connector holes and three screw holes.  
3. Make an interlock circuit as shown in “To Make Interlock Circuit” on page 2-15.  
4. Make cable connections to the DUT interface. Refer to “To Make Connections  
to DUT Interface” on page 2-13.  
2-18  
Agilent B2200 User’s Guide, Edition 2  
 
Installation  
Output Connections  
Table 2-5  
Recommended Parts  
Description  
Agilent Part No.  
Interlock Connector (6 pin, female)  
Switch  
1252-1419C  
3101-0302 or 3101-3241  
1450-0641  
LED (VF @ 2.1 V @ IF = 10 mA)  
Wire  
8150-5680  
1250-2457  
8121-1191  
Triaxial Connector (female)  
Low Noise Coaxial Cable  
Table 2-6  
Dimensions of Connector Holes  
Triaxial Connector (in mm)  
Interlock Connector (in mm)  
11.3  
1.8  
8.2  
Kelvin Triaxial Connector (in mm)  
14  
11  
11  
14  
2.8  
10.3  
2 − ∅11.3  
3.2  
2 M3 x 0.5  
Agilent B2200 User’s Guide, Edition 2  
2-19  
   
Installation  
Input Connections  
Input Connections  
This section explains how to connect instruments to the Agilent B2200.  
WARNING  
Turn off all instruments that will be connected to the Agilent B2200. And do  
not turn them on until the connection described in this section is completed. If  
you ignore this warning, you may be exposed to dangerous voltage.  
The Agilent B2200 input connector has eight SMU input connectors and six AUX  
input connectors. Table 2-7 shows cables used to connect the instrument for typical  
applications using the Agilent B2200.  
You can make the Kelvin connections when using the E5260/E5270/4142/4156  
SMU or 41501 HPSMU. Then use cables as follows:  
Agilent 16493K Kelvin triaxial cable for Agilent E5260/E5270/4156/41501  
Agilent 16494B Kelvin triaxial cable for Agilent 4142B  
To connect the Kelvin triaxial cable to the Agilent B2200 input, use pairs (couple  
ports) of the SMU input connectors (1&2, 3&4, 5&6, or 7&8). The input signal for  
each Kelvin input is as follows:  
Odd input port number: Force  
Even input port number: Sense  
Ground Unit (GNDU) can sink over 1 A. For the measurement over 1 A, connect the  
GNDU to the connector plate directly by using the Agilent 16493H GNDU cable.  
If you never perform the measurement over 1 A, you can connect the GNDU to the  
Agilent B2200 inputs. Then use the Agilent 16493N GNDU cable. This cable  
requires the Kelvin connection.  
CAUTION  
Make sure current to the GNDU is less than 1 A if you connect GNDU to the  
Agilent B2200 inputs. The maximum input current of the Agilent B2200 is 1A.  
2-20  
Agilent B2200 User’s Guide, Edition 2  
 
Installation  
Input Connections  
Table 2-7  
B2200 Input Connections  
Instrument  
B2200  
Application  
Input  
Connector  
Cable Required  
Agilent  
Model No. Connector  
Output  
DC  
Measurement  
4155  
4156  
MPSMU  
VSU  
SMU1 to 8 16494A triaxial cable  
AUX1 to 6 BNC cable  
VMU  
Intlk  
AUX1 to 6 BNC cable  
16493J interlock cable  
(for direct connection to connector plate)  
HRSMU  
SMU1 to 8 16493K Kelvin triaxial cable or  
16494A triaxial cable  
VSU  
VMU  
Intlk  
AUX1 to 6 BNC cable  
AUX1 to 6 BNC cable  
16493J interlock cable  
(for direct connection to connector plate)  
41501  
HPSMU  
SMU1 to 8 16493K Kelvin triaxial cable or  
16494A triaxial cable  
MPSMU  
PGU  
SMU1 to 8 16494A triaxial cable  
AUX1 to 6 BNC cable  
GNDU  
SMU1 to 8 16493N GNDU Kelvin cable  
16493H GNDU cable  
(for direct connection to connector plate)  
4142B  
SMU  
SMU1 to 8 16494B Kelvin triaxial cable or  
16494A triaxial cable  
VSU  
VMU  
GNDU  
AUX1 to 6 BNC cable  
AUX1 to 6 BNC cable  
SMU1 to 8 16493HN GNDU Kelvin cable  
16493H GNDU cable  
(for direct connection to connector plate)  
Intlk  
16435A interlock cable adapter  
and 16493J interlock cable  
(for direct connection to Connector Plate)  
Agilent B2200 User’s Guide, Edition 2  
2-21  
 
Installation  
Input Connections  
Instrument  
B2200  
Input  
Connector  
Application  
Cable Required  
Agilent  
Output  
Model No. Connector  
DC  
E5260/  
SMU  
SMU1 to 8 16493K Kelvin triaxial cable or  
16494A triaxial cable  
Measurement E5270  
GNDU  
SMU1 to 8 16493N GNDU Kelvin cable  
16493H GNDU cable  
(for direct connection to connector plate)  
Interlock  
16493J interlock cable  
(for direct connection to connector plate)  
C
C Meter  
High, Low CMH/CML 16494F CMU cable for 4 terminal pairs  
Measurement  
Pulse Input  
Bias Input  
Pulse  
Generator  
Output  
Output  
AUX1 to 6 BNC cable  
AUX1 to 6 BNC cable  
Power  
Supply  
In the table above,  
CMH/CML connectors are a pair of the AUX inputs.  
The Agilent B2200 does not have the interlock connector. Connect directly from  
instrument to the connector plate which has an interlock connector.  
The Agilent 41422A/41423A (HCU/HVU) of the Agilent 4142B cannot be used  
with the Agilent B2200.  
To connect instrument output that uses banana plug to the AUX connector, use  
the dual banana plug to BNC adapter (Agilent part number 1251-2277).  
2-22  
Agilent B2200 User’s Guide, Edition 2  
Installation  
Measurement Cable Length  
Measurement Cable Length  
This section describes how to calculate the total guard capacitance when using an  
SMU (source monitor unit). When using an SMU, the length of measurement cables  
is limited by the guard capacitance of the cables. The guard capacitance means the  
capacitance between the signal line (Force or Sense) and the Guard line.  
Table 2-8 lists the guard capacitance for each element of the Agilent B2200’s  
measurement environment.  
When using the 4155/4156/4142B/E5260/E5270, the maximum limit of the guard  
capacitance is approximately 900 pF. So, you add the following and total must be  
less than 900 pF:  
Guard capacitance of cable from SMU to B2200 inputs.  
Guard capacitance of mainframe and modules.  
Guard capacitance of cable from B2200 outputs.  
Guard capacitance of cable from connector plate to DUT.  
Other capacitances, such as for probe card.  
The following is an example to calculate the guard capacitance of a measurement  
environment:  
Installed Plug-in cards: Four B2210As (145 pF + 8 pF × 3)  
Input Cable:  
16494A-003 (75 pF)  
16494A-001 (130 pF)  
Output Cable:  
Cable from Connector  
Plate to DUT:  
8121-1191, 1 m (130 pF)  
Probe Card:  
Guard Capacitance approximately 10 pF (example)  
In this environment, the total guard capacitance (Cg-total) is as follows.  
Cg-total = 145+(8×3)+75+130+130+10 pF = 514 pF < 900 pF  
Even if you use other instruments, you can use the table for reference about the  
measurement cable length and capacitance.  
Agilent B2200 User’s Guide, Edition 2  
2-23  
 
Installation  
Measurement Cable Length  
Table 2-8  
Connection  
Guard Capacitances of B2200 Measurement Environment (Typical)  
Cable  
Guard  
Agilent Model/Part No.  
Length  
Capacitance  
from SMU to B2200 16494A-003 (for non-Kelvin)  
80 cm  
1.5 m  
80 cm  
1.5 m  
75 pF  
130 pF  
inputs  
16494A-001 (for non-Kelvin)  
16494B-003 (for Kelvin, 4142B)  
16493K-001 (for Kelvin)  
90 pF  
150 pF  
mainframe and  
modules  
B2210A (only 1 card is installed)  
B2210A (more 1 card)  
145 pF  
8 pF/extra card  
145 pF  
B2211A (only 1 card is installed)  
B2211A (more 1 card)  
8 pF/extra card  
130 pF  
from B2200 outputs 16494A-001 (for non-Kelvin)  
to connector plate or  
1.5 m  
3 m  
4 m  
1.5 m  
3 m  
4 m  
X m  
16494A-002 (for non-Kelvin)  
probe card interface  
240 pF  
16494A-003 (for non-Kelvin)  
16494B/C-001 (for Kelvin)  
16494B/C-002 (for Kelvin)  
16494C-003 (for Kelvin)  
340 pF  
140 pF  
260 pF  
340 pF  
from connector  
plate to DUT  
Agilent Part No. 8121-1191  
130 pF per m  
probe card interface B2220A  
70 pF  
2-24  
Agilent B2200 User’s Guide, Edition 2  
 
Installation  
Maintenance  
Maintenance  
Maintenance should be performed periodically to keep the B2200 in good condition.  
Calibration  
Calibration must be performed periodically so that the instruments satisfy the  
specifications, and keep a good condition. It is recommended to perform a  
calibration once a year at least. For calibration, contact your nearest Agilent  
Technologies Service Center. Trained service personnel will perform calibration  
(performance verification).  
Cleaning  
Before performing cleaning, turn off the instrument, and disconnect power cable  
from the rear panel. Use a dry cloth to clean the external case parts.  
To prevent electrical shock, do not perform cleaning when the instrument is turned  
on, and do not use a wet cloth.  
Agilent B2200 User’s Guide, Edition 2  
2-25  
     
Installation  
Maintenance  
2-26  
Agilent B2200 User’s Guide, Edition 2  
3
Front Panel Operation  
 
Front Panel Operation  
This chapter explains the front panel operation and the switch control functions of  
the Agilent B2200, also provides the reference information of the front panel keys  
and display.  
“Operation”  
“Switch Control Functions”  
“Display Functions”  
“Front Panel Keys”  
“Setup Menus”  
3-2  
Agilent B2200 User’s Guide, Edition 2  
Front Panel Operation  
Operation  
This section describes operations of the Agilent B2200.  
“To Change Channel Configuration Mode”  
“To Change Connection Sequence”  
“To Set Beeper”  
“To Set GPIB Address”  
“To Set Remote Display Mode”  
“To Return to Local Mode”  
NOTE  
About moving cursor, selecting value, and changing display  
Use the right/left arrow key to move the cursor to the desired function name. Use the  
up/down arrow key to select the value for the setup item.  
The Enter key makes the setup effective, and returns the display to the previous  
menu or display.  
The Exit key returns the display to the previous menu or display without making the  
setup effective.  
Agilent B2200 User’s Guide, Edition 2  
3-3  
 
Front Panel Operation  
Operation  
To Initialize Agilent B2200  
1. Press the Shift key and the Local key. Then the LCD will display NO.  
2. Press the arrow key to displays YES, then press the Enter key to initialize the  
Agilent B2200, or press the Exit key to cancel initialization.  
To Enable Light Pen  
Turn the Agilent B2200 off, and connect the light pen to the Light Pen connector at  
the right down corner of the front panel. After that, perform the following procedure  
to enable the light pen.  
1. Press the Menu key. The setup menu will be displayed on the LCD.  
2. Press the arrow key to move the cursor to PEN, then press the Enter key.  
3. Press the arrow key to display ENABLE, then press the Enter key to enable the  
light pen, or press the Exit key to cancel changing the setup.  
NOTE  
Light Pen  
You can control the Agilent B2200 functions and the switch connections by using  
the front panel keys. And the Agilent B2200 allows you to use the Agilent 16443A  
Light Pen to control the switch connections.  
Pointing a LED on the LED matrix by using the light pen controls the setting of the  
switch specified by the LED.  
To Change Channel Configuration Mode  
1. Press the Shift key and the Bias Mode key. The LCD displays the present setting  
(AUTO or NORMAL).  
2. Press the arrow key to select the mode, then press the Enter key. The LCD  
displays NO.  
3. Press the arrow key to display YES, then press the Enter key. Or press the Exit  
key to cancel this operation.  
Changing the channel configuration mode resets the Agilent B2200.  
3-4  
Agilent B2200 User’s Guide, Edition 2  
     
Front Panel Operation  
Operation  
To Change Connection Rule  
1. Press the Rule key. The LCD displays the present setting (FREE or SINGLE).  
2. Press the arrow key to select the connection rule, then press the Enter key. Or  
press the Exit key to cancel this operation.  
To Change Connection Sequence  
1. Press the Shift key and the Rule key. The LCD displays the present setting  
(Break_Before_Make, Make_Before_Break, or No_Sequence).  
2. Press the arrow key to select the connection sequence, then press the Enter key.  
Or press the Exit key to cancel this operation.  
To Control Switch Condition  
If you use the light pen, you do not need the following instruction to control the  
switch condition.  
1. If you are NOT watching the orange-blink-LED on the LED matrix, press the  
Open/Close key. The blink-LED will appear.  
2. Press the arrow keys to move the blink-point.  
3. Press the Open/Close key to open/close the switch specified by the blink-LED.  
The green-LEDs indicate the switches that are closed now.  
4. Repeat 2 and 3 until the switch setup is completed.  
Blinking of LED will be stopped by pressing the front panel key other than the  
Open/Close, Fast, or arrow keys.  
NOTE  
When Kelvin cables are connected to Outputs  
For the Kelvin outputs, the input-output paths must be controlled as shown below.  
For the input couple port:  
Control the switches to connect the input couple port to the Kelvin output.  
For the single input port:  
Control the switches to connect the input port to two ports of the Kelvin output.  
If you do not comply with this note, leakage current between the Kelvin paths will  
cause measurement error.  
Agilent B2200 User’s Guide, Edition 2  
3-5  
     
Front Panel Operation  
Operation  
To Open All Switches  
1. Press the Open All key.  
2. Press the arrow key to display YES, then press the Enter key.  
NOTE  
Status after this operation  
When the bias mode is ON, the bias-enabled output ports will be connected to the  
input bias port.  
When the ground mode is ON, the ground-enabled input/output ports will be  
connected to the input ground port.  
To Save/Load Setup Data  
The Agilent B2200 provides the internal memory used to save the setup condition.  
Eight setup conditions can be saved into the memory.  
To save the setup  
1. Press the Shift key and the Load Memory key.  
2. Press the arrow key to select the internal memory (01 to 08).  
3. Press the Enter key to save the setup. Or press the Exit key to cancel this  
operation.  
Wait until the save operation is completed. Then, the LCD displays the status  
and setup conditions.  
To load the setup  
1. Press the Load Memory key.  
2. Press the arrow key to select the internal memory (01 to 08).  
3. Press the Enter key to load the setup. Or press the Exit key to cancel this  
operation. Wait until the load operation is completed. Then, the LCD displays  
the status and setup conditions.  
3-6  
Agilent B2200 User’s Guide, Edition 2  
   
Front Panel Operation  
Operation  
To Use Bias Mode  
Bias mode cannot be set to ON when the ground mode is ON.  
1. Change the bias-enabled output ports.  
a. Press the Shift key and the Port Function key.  
b. Move the cursor to SET, then press the Enter key.  
c. Move the cursor to DISABLE, then press the Enter key.  
The LCD displays one of the bias-enabled output port numbers which can be  
changed to the bias-disabled. Or if there is no bias-enabled output port, the  
LCD displays Channel: No Channel. Press the arrow key to select the  
output port to be changed to the bias-disabled, then press the Enter key.  
d. Move the cursor to ENABLE, then press the Enter key.  
The LCD displays one of the bias-disabled output port numbers which can  
be changed to the bias-enabled. Or if there is no bias-disabled output ports,  
the LCD displays Channel: No Channel. Press the arrow key to select  
the output port to be changed to the bias-enabled, then press the Enter key.  
e. Repeat c and d until the bias-enabled port setup is completed. After that,  
press the Exit key more than once to display the status and setup conditions.  
2. Press the Bias Mode key to set the bias mode ON. Pressing the key again sets the  
mode OFF.  
3. Change the bias port (default setting: 10).  
a. When the bias mode is ON, press the Port Function key. The blink-cursor  
appears on the first line of the LCD.  
b. Press the right/left arrow key to move the cursor on the input port number  
(01 to 14) that will be set to the input bias port.  
c. Press the up/down key to display B, then press the Enter key.  
Agilent B2200 User’s Guide, Edition 2  
3-7  
   
Front Panel Operation  
Operation  
To Use Ground Mode  
Ground mode cannot be set to ON when the bias mode is ON.  
1. Change the ground-enabled output ports.  
a. Press the Shift key and the Open All key.  
b. Move the cursor to SET, then press the Enter key.  
c. Move the cursor to ENABLE, then press the Enter key.  
The LCD displays one of the ground-disabled output port numbers which  
can be changed to the ground-enabled. Or if there is no ground-disabled  
output ports, the LCD displays Channel: No Channel. Press the arrow  
key to select the output port to be changed to the ground-enabled, then press  
the Enter key.  
d. Move the cursor to DISABLE, then press the Enter key.  
The LCD displays one of the ground-enabled output port numbers which can  
be changed to the ground-disabled. Or if there is no ground-enabled output  
port, the LCD displays Channel: No Channel. Press the arrow key to  
select the port to be changed to the ground-disabled, then press the Enter key.  
e. Repeat c and d until the ground-enabled port setup is completed. After that,  
press the Exit key more than once to display the status and setup conditions.  
2. Press the Shift key and the Couple Mode key to set the ground mode ON.  
Performing this operation again sets the mode OFF.  
3. Change the ground port (default setting: 12).  
a. When the ground mode is ON, press the Port Function key. The blink-cursor  
appears on the first line of the LCD.  
b. Press the right/left arrow key to move the cursor on the input port number  
(01 to 14) that will be set to the input ground port.  
c. Press the up/down key to display G.  
4. Change the ground-enabled input ports (multiple ports can be set). Repeat a and  
b for the all ground-enabled input ports, then press the Enter key.  
a. Press the right/left arrow key to move the cursor on the input port number  
(01 to 08) that will be set to the ground-enabled input port.  
b. Press the up/down key to display -.  
Open the ground-enabled input ports to prevent the instrument from damage.  
3-8  
Agilent B2200 User’s Guide, Edition 2  
 
Front Panel Operation  
Operation  
To Use Couple Mode  
1. Press the Couple Mode key to set the couple mode ON. Pressing the key again  
sets the mode OFF.  
2. Perform the input couple port detection.  
a. Press the Menu key. The setup menu will be displayed on the LCD.  
b. Move the cursor to SCAN, then press the Enter key. The LCD displays NO.  
c. Press the arrow key to display YES, then press the Enter key.  
d. Press the Exit key.  
3. Change the couple ports (multiple ports can be set).  
a. When the couple mode is ON, press the Port Function key. The blink-cursor  
appears on the first line of the LCD.  
b. Press the right/left arrow key to move the cursor on the input port number  
(01 to 14) that will be set to the input couple port.  
c. Press the up/down key to display C.  
d. Repeat b and c for the all couple ports, then press the Enter key.  
To Display Firmware Revision  
1. Press the Menu key. The setup menu will be displayed on the LCD.  
2. Move the cursor to CONFIG, then press the Enter key.  
3. Move the cursor to REVISION, then press the Enter key. The firmware revision  
is displayed.  
To Display Module Information  
1. Press the Menu key. The setup menu will be displayed on the LCD.  
2. Move the cursor to CONFIG, then press the Enter key.  
3. Move the cursor to UNIT, then press the Enter key. The module information is  
displayed.  
4. Press the arrow keys to display another module information item.  
Agilent B2200 User’s Guide, Edition 2  
3-9  
     
Front Panel Operation  
Operation  
To Read Error Message  
1. Press the Menu key. The setup menu will be displayed on the LCD.  
2. Move the cursor to ERROR, then press the Enter key.  
3. Move the cursor to DISPLAY, then press the Enter key to display the message.  
4. Press the arrow keys to read another error message. A maximum of four error  
messages can be stored.  
To clear error buffer  
1. Press the Menu key. The setup menu will be displayed on the LCD.  
2. Move the cursor to ERROR, then press the Enter key.  
3. Move the cursor to CLEAR, then press the Enter key. The LCD displays NO.  
4. Press the arrow key to display YES, then press the Enter key to clear the error  
buffer, or press the Exit key to cancel operation.  
To Set Beeper  
1. Press the Menu key. The setup menu will be displayed on the LCD.  
2. Move the cursor to BEEP, then press the Enter key.  
3. Press the arrow keys to select the beep ON or OFF.  
4. Press the Enter key to make the setup effective, or press the Exit key to cancel  
changing the setup.  
To Set GPIB Address  
1. Press the Menu key. The setup menu will be displayed on the LCD.  
2. Move the cursor to CONFIG, then press the Enter key.  
3. Move the cursor to ADDRESS, then press the Enter key.  
4. Press the arrow keys to set the GPIB address.  
5. Press the Enter key to make the setup effective, or press the Exit key to cancel  
changing the setup. To be effective the new address, reboot the Agilent B2200.  
3-10  
Agilent B2200 User’s Guide, Edition 2  
     
Front Panel Operation  
Operation  
To Set Remote Display Mode  
This instruction enables or disables the data display in the GPIB remote condition.  
See “RMT_DSPL” on page 3-33.  
LCD  
1. Press the Menu key. The setup menu will be displayed on the LCD.  
2. Move the cursor to RMT_DSPL, then press the Enter key.  
3. Move the cursor to LCD, then press the Enter key.  
4. Press the arrow keys to set the remote display mode ON or OFF.  
5. Press the Enter key to make the setup effective, or press the Exit key to cancel  
changing the setup.  
LED matrix  
1. Press the Menu key. The setup menu will be displayed on the LCD.  
2. Move the cursor to RMT_DSPL, then press the Enter key.  
3. Move the cursor to LED, then press the Enter key.  
4. Press the arrow keys to set the remote display mode ON or OFF.  
5. Press the Enter key to make the setup effective, or press the Exit key to cancel  
changing the setup.  
To Return to Local Mode  
Press the Local key. If the front panel keys are locked, send the :SYST:KLC  
command from an external computer, then press the Local key.  
Agilent B2200 User’s Guide, Edition 2  
3-11  
   
Front Panel Operation  
Switch Control Functions  
This section introduces the switch control functions of the Agilent B2200.  
“Connection Sequence”  
“Bias Mode”  
“Ground Mode”  
“Couple Mode”  
3-12  
Agilent B2200 User’s Guide, Edition 2  
 
Front Panel Operation  
Switch Control Functions  
Channel Configuration Mode  
The Agilent B2200 provides the two channel configuration modes, Normal and  
Auto. The configuration mode defines the way to control multiple switch modules  
installed in the mainframe.  
Normal  
Each module is considered independently. So each module is  
always a 12 output switching matrix.  
The switch control functions need to be defined for each  
module.  
Auto  
Default mode at Power On or *RST command.  
To use this mode, the same type of module must be installed in  
the mainframe from slot number 1 continuously (slots 1 to 2; or  
slots 1 to 3; or slots 1 to 4).  
The installed modules are treated as one switching matrix. For  
example, if the modules are installed in slots 1 to 3, the Agilent  
B2200 works as a 36 output switching matrix.  
The switch control functions are available for the switching  
matrix.  
Agilent B2200 User’s Guide, Edition 2  
3-13  
 
Front Panel Operation  
Switch Control Functions  
Connection Rule  
Connection rule is used to specify if an input or output port can have multiple  
connections. For each module installed in the mainframe (for auto configuration  
mode, installed modules are treated as one module), you can specify one of the  
following connection rules:  
Single:  
Each input port can be connected to only one output port, and  
each output port can be connected to only one input port. So,  
existing connection to a port will be disconnected when a new  
connection is made.  
If normal configuration mode is set to the Agilent B2200 with  
multiple modules, an input/output path is effective for each  
module. For example, if normal configuration mode is set to the  
Agilent B2200 with two modules, and single rule is set to both  
modules, the following connection is possible.  
1
2
Input  
3
14  
2
3
2
3
12  
12  
1
1
Close  
Open  
Module 1 Output  
Module 2 Output  
Free:  
Each input port can be connected to multiple output ports, and  
each output port can be connected to multiple input ports.  
CAUTION  
If the Free connection rule has been specified, ensure multiple input ports are not  
connected to the same output port. Such configurations can cause damage.  
3-14  
Agilent B2200 User’s Guide, Edition 2  
 
Front Panel Operation  
Switch Control Functions  
Connection Sequence  
NOTE  
This is only for modules that are set to the single connection rule. Refer to  
“Connection Rule” on page 3-14.  
Connection sequence specifies the open/close sequence of the relays when changing  
from an existing connection to a new connection.  
You can select one of the three connection sequences shown in Table 3-1 for each  
module installed in the Agilent B2200.  
When an existing connection is changed to a new connection, the relays are  
opened/closed as follows:  
Table 3-1  
Connection Sequence  
Connection  
Relay Operation  
Sequence  
Break Before Make 1. Disconnect previous route.  
2. Wait for relays to open.  
3. Connect new route.  
Make Before Break 1. Connect new route.  
2. Wait for relays to close.  
3. Disconnect previous route.  
No Sequence  
1. Disconnect previous route.  
2. Connect new route.  
Agilent B2200 User’s Guide, Edition 2  
3-15  
   
Front Panel Operation  
Switch Control Functions  
Bias Mode  
Bias mode is useful for connecting the same input to multiple channels (output  
ports) simultaneously. And the connections will be kept until the other input port  
connection will be changed.  
When the bias mode is ON, the input bias port is connected to all bias enabled  
output ports that are not connected to any other input ports. You cannot directly  
control which output ports are connected to the input bias port:  
If another input port is disconnected from a bias enabled output port, the output  
port is automatically connected to the input bias port.  
If another input port is connected to a bias enabled output port, the output port is  
automatically disconnected from the input bias port.  
Bias disabled output ports are never connected to the input bias port when the  
bias mode is ON.  
You can independently set the bias port and control the bias mode for each  
module when the configuration mode is normal.  
When the bias mode is OFF, the input bias port is the same as the other input ports,  
so relays can be controlled directly to connect to output ports.  
Figure 3-1 shows an example. Input 2 is the input bias port. When the bias mode is  
turned ON, all bias enabled output ports (that are not connected to other input ports)  
are connected to the input bias port. So, the output ports that are connected to input  
port 1 and 3 are not connected to the input bias port.  
Figure 3-1  
Example of Bias Mode  
Input 01  
Input 02  
Input 03  
Bias port  
Input 14  
Output port  
Open  
Close  
Connects to bias port  
Connects to input 03  
Connects to input 01  
3-16  
Agilent B2200 User’s Guide, Edition 2  
   
Front Panel Operation  
Switch Control Functions  
NOTE  
Connection rule cannot be specified for the input bias port, which can always be  
connected to multiple output ports.  
Connection sequence (to connect input bias port to output ports) is always  
Break-Before-Make.  
Bias mode cannot be set to ON when the ground mode is ON.  
If the bias input port and a couple port have been assigned to the same input port, the  
bias mode and the couple mode cannot be used in parallel.  
Agilent B2200 User’s Guide, Edition 2  
3-17  
Front Panel Operation  
Switch Control Functions  
Ground Mode  
Ground mode operation is similar to the bias mode operation. The ground mode is  
useful for connecting the same input to multiple channels (output ports) and the  
unused input ports simultaneously. And the connections will be kept until the other  
input port connection will be changed. The ground mode is especially useful for  
settling the potential of the unused input/output paths.  
If you assign the input port 12 (AUX Input 12) as the ground port, open the input  
port 12. This ground port will be internally connected to the ground when the  
ground mode is set to ON.  
When the ground mode is ON, the input ground port is connected to all ground  
enabled input ports (unused ports) and the ground enabled output ports that are not  
connected to any other input ports. You cannot directly control which output ports  
are connected to the input ground port:  
If another input port is disconnected from a ground enabled output port, the  
output port is automatically connected to the input ground port.  
If another input port is connected to a ground enabled output port, the output  
port is automatically disconnected from the input ground port.  
Ground disabled output ports are never connected to the input ground port when  
the ground mode is ON.  
You can independently set the ground port and control the ground mode for each  
module when the configuration mode is normal.  
When the ground mode is OFF, the input ground port is the same as the other input  
ports, so relays can be controlled directly to connect to output ports.  
NOTE  
Connection rule cannot be specified for the input ground port, which can always be  
connected to multiple output ports.  
Connection sequence (to connect input ground port to output ports) is always  
Break-Before-Make.  
Ground mode cannot be set to ON when the bias mode is ON.  
The input ground port and a ground enabled input port cannot be assigned to the  
same input port.  
If the ground input port and a couple port have been assigned to the same input port,  
the ground mode and the couple mode cannot be used in parallel.  
If a ground enabled input port and a couple port have been assigned to the same  
input port, the ground mode and the couple mode cannot be used in parallel.  
3-18  
Agilent B2200 User’s Guide, Edition 2  
 
Front Panel Operation  
Switch Control Functions  
CAUTION  
Opening the ground enabled input ports  
Open the ground enabled input ports (unused input ports). If any equipment is  
connected to the ground enabled input ports, turning the ground mode ON may  
cause damage in the equipment.  
Agilent B2200 User’s Guide, Edition 2  
3-19  
Front Panel Operation  
Switch Control Functions  
Couple Mode  
Couple mode is useful for making the Kelvin connections. When the couple mode is  
ON, the input couple ports will be connected to the output couple ports. For the  
available couple ports, see Table 3-2.  
For the couple port connections, the odd/even number of the input port will be  
connected to the odd/even number of the output port respectively. Then the even  
number is always the odd number plus 1.  
You can independently specify the same or different couple ports and control the  
couple mode for each module when the configuration mode is normal.  
Table 3-2  
Input Couple Port Numbers and Output Couple Port Numbers  
Available input  
couple port No.  
Input ports  
Available output couple port No.  
1
3
SMU 1, 2  
SMU 3, 4  
SMU 5, 6  
SMU 7, 8  
AUX 9, 10  
AUX 11, 12  
1 and 2, 3 and 4, . . . , 11 and 12  
(for normal configuration mode)  
1 and 2, 3 and 4, . . . , 47 and 48  
(for auto configuration mode)  
5
7
9
11  
13  
AUX 13, 14  
(CMH, CML)  
For example, if you specify the SMU input 1 for the couple port and set the couple  
mode ON, the input 1 will be connected to an odd number of the output port and the  
input 2 will connected to the next number of the output port. And if you specify to  
connect the input 1 to the output 10, the input 1 will be connected to the output 9 and  
the input 2 will be connected to the output 10.  
NOTE  
If the bias input port and a couple port have been assigned to the same input port, the  
bias mode and the couple mode cannot be used in parallel.  
If the ground input port and a couple port have been assigned to the same input port,  
the ground mode and the couple mode cannot be used in parallel.  
If a ground enabled input port and a couple port have been assigned to the same  
input port, the ground mode and the couple mode cannot be used in parallel.  
3-20  
Agilent B2200 User’s Guide, Edition 2  
   
Front Panel Operation  
Switch Control Functions  
NOTE  
Couple Port Detection Function  
The Agilent B2200 provides the function to detect the input ports that connect the  
Kelvin triaxial cable and set the ports as the couple port automatically. See “SCAN”  
on page 3-32  
Agilent B2200 User’s Guide, Edition 2  
3-21  
Front Panel Operation  
Display Functions  
Display Functions  
The Agilent B2200 provides LED matrix, LCD, and 18 front panel keys for front  
panel operation. This section explains the display of the LED matrix and the LCD.  
“LED Matrix”  
“LCD”  
LED Matrix  
Agilent B2200 has four blocks of 14 × 12 LED matrix. They display the switch  
condition of the switch modules installed in the slot 1 to slot 4. See Table 3-3.  
Also LEDs labeled Card 1 to Card 4 are located above the LED matrices. They  
indicate the status of the module installed in the slot 1 to 4, respectively. See Table  
3-4.  
Table 3-3  
LED Matrix  
LED color  
Explanation  
Green  
The switch has been closed to connect the output channel to the  
input port.  
Red  
Orange  
-
The switch has been closed to connect the output channel to the  
input bias port or the input ground port.  
The LED also blinks. The switch condition can be changed by the  
Open/Close key.  
The LED lights out. The switch has been opened.  
Table 3-4  
Card 1/2/3/4 LED  
LED color  
Explanation  
Red  
Green  
-
The switch module has failed the selftest or the diagnostics.  
The switch module has operated normally.  
The LED lights out. No switch module has been installed in this  
slot.  
3-22  
Agilent B2200 User’s Guide, Edition 2  
       
Front Panel Operation  
Display Functions  
LCD  
The LCD displays the setup information and the status information as shown in  
Table 3-5. Display example is shown in Figure 3-2.  
Figure 3-2  
LCD Display Example  
01  
02  
03  
04  
05  
06  
07  
08  
09  
10  
11  
12  
13  
14  
Shift  
- C C -  
A
G
BBM  
FREE  
Sequence  
Lock Error  
Bias  
Gnd  
The first line shows the input port status, couple port, bias port, ground port, or  
ground enabled port.  
The second line shows the instrument status and the mode status.  
The example of Figure 3-2 notifies you the following information.  
Input 5 and 6 are a couple port.  
Input 12 is the input ground port.  
Input 4 and 7 are the ground enabled input port.  
Channel configuration mode is auto.  
Couple mode is ON.  
Ground mode is ON.  
Connection rule is free.  
Connection sequence is break before make.  
Agilent B2200 User’s Guide, Edition 2  
3-23  
   
Front Panel Operation  
Display Functions  
Table 3-5  
LCD Display Items  
Label  
Description  
01 to 14  
Port function assigned to the input ports 1 to 14. B(bias port), C(couple port),  
G(ground port), -(ground enabled port), or blank (no function).  
Shift  
Shift key status indicator. The triangle mark appears when the sub key is active.  
Remote  
Remote status indicator. The triangle mark appears when the Agilent B2200 is in  
the GPIB remote condition.  
Lock  
Key status indicator. The triangle mark appears when the front panel keys are  
locked by an external computer.  
Error  
Error status indicator. The triangle mark appears when any error has occurred.  
Card#  
Slot number of the switch module to be monitored. Displays Awhen the channel  
configuration mode is Auto. In the auto mode, the modules installed in the  
mainframe from slot number 1 continuously will be controlled as one module. Or  
displays 1to 4when the channel configuration mode is Normal. The module can  
be selected by the Card Selection key.  
Bias  
Couple  
Gnd  
Bias mode status indicator. The triangle mark appears when the mode is ON.  
Couple mode status indicator. The triangle mark appears when the mode is ON.  
Ground mode status indicator. The triangle mark appears when the mode is ON.  
Connection rule, FREE(plural route mode) or SINGLE(single route mode).  
Rule  
Sequence Connection sequence, BBM(Break Before Make), MBBR(Make Before Break), or  
NO_SEQ(No Sequence).  
NOTE  
In the remote mode and the default setting, only the Remote, Lock, and Error  
indicators are available. To display all information, set the RMT_DSPL function  
ON. Refer to “RMT_DSPL” on page 3-33.  
3-24  
Agilent B2200 User’s Guide, Edition 2  
 
Front Panel Operation  
Front Panel Keys  
Front Panel Keys  
The Agilent B2200 provides LED matrix, LCD, and 18 front panel keys for front  
panel operation. The front panel keys are used to change the instrument settings, the  
switch conditions, and so on.  
Function  
Edit  
Local  
Sets the Agilent B2200 to the local condition.  
Used to reset the Agilent B2200.  
Reset  
(Shift+Local)  
Displays the following message. Press the arrow key to set YES, and then press the  
Enter key to reset the instrument.  
NO  
Menu  
Displays the setup menu. See “Setup Menu” on page 3-30.  
Selftest  
Displays the selftest menu. See “Selftest Menu” on page 3-34.  
(Shift+Menu)  
Agilent B2200 User’s Guide, Edition 2  
3-25  
 
Front Panel Operation  
Front Panel Keys  
Function key group  
The Function key group is used to change the input ports and modes.  
Function  
Edit  
Port Function  
Enters into the port function selection mode. In this mode, the arrow keys, Fast key,  
Exit key, and Enter key are available. Other keys will work to exit this mode.  
In this mode, move cursor right or left to specify the input port, and press the up or  
down arrow key to select the port function, B (bias port), C (couple port), G (ground  
port), or -(ground enable port). The port function can be selected when the  
corresponding port operation mode (bias mode, couple mode, or ground mode) is set  
to ON. The bias mode and the ground mode cannot be set to ON simultaneously.  
Bias Ch (Shift+  
Port Function)  
Displays the following menu.  
SET VIEW  
SETdisplays the following menu. The following functions are used to set the  
bias-enabled or bias-disabled channel (output port).  
ENABLE DISABLE  
The function displays the following message for example. And then press Enter  
to bias-enable or bias-disable the channel.  
Channel: 01  
If there is no enabled or disabled channel, the following message will appear.  
Channel: No Channel  
VIEWdisplays the following message for example, and is used to see the bias  
status of the channel.  
Channel 01: ENABLED  
3-26  
Agilent B2200 User’s Guide, Edition 2  
 
Front Panel Operation  
Front Panel Keys  
Bias Mode  
Sets the bias mode ON or OFF.  
When the bias mode is set to ON, the LCD shows which input port is the bias port,  
and the input bias port will be connected to the bias-enabled channels that are not  
connected to other input port. And LEDs will be turned red to show the bias port  
connections.  
Ch Mode (Shift+  
Bias Mode)  
Sets the channel configuration mode AUTO or NORMAL.  
Changing the mode will reset the Agilent B2200 settings.  
Couple Mode  
Sets the couple mode ON or OFF.  
When the couple mode is set to ON, the LCD shows which input ports are the  
couple port.  
Gnd Mode (Shift+ Sets the ground mode ON or OFF.  
Couple Mode)  
When the ground mode is set to ON, the LCD shows which input port is the ground  
port, and the input ground port will be connected to the ground-enabled channels  
that are not connected to other input port. And LEDs will be turned red to show the  
ground port connections. The ground enabled input ports are also connected to the  
input ground port. Nothing must be connected to the ground enabled input ports.  
Rule  
Displays the following message. Selects the connection rule FREE (plural route  
connections) or SINGLE (single route connection).  
FREE  
Sequence  
(Shift+Rule)  
Displays the following message. Selects the connection sequence  
Break_Before_Make, Make_Before_Break, or No_Sequence.  
Break_Before_Make  
Load Memory  
Displays the following message. Used to select the Agilent B2200 setup data stored  
in the internal memory, and loads the data.  
Settings: 01 [  
]
Save Memory  
(Shift+  
Load Memory)  
Displays the following message. Used to select the internal memory to store the  
Agilent B2200 setup data, and stores the data. Up to eight setup data can be saved.  
Settings: 01 [  
]
Open All  
Displays the following message. Press the arrow key to set YES, and then press the  
Enter key to open all relays.  
NO  
Agilent B2200 User’s Guide, Edition 2  
3-27  
Front Panel Operation  
Front Panel Keys  
Gnd Ch  
Displays the following menu.  
(Shift+Open All)  
SET VIEW  
SETdisplays the following menu. The following functions are used to set the  
ground-enabled or ground-disabled channel (output port).  
ENABLE DISABLE  
The function displays the following message for example. And then press Enter  
to ground-enable or ground-disable the channel.  
Channel: 01  
If there is no enabled or disabled channel, the following message will appear.  
Channel: No Channel  
VIEWdisplays the following message for example, and is used to see the ground  
status of the channel.  
Channel 01: ENABLED  
3-28  
Agilent B2200 User’s Guide, Edition 2  
Front Panel Operation  
Front Panel Keys  
Edit key group  
The Edit key group is used to select the switch module to be monitored, move the  
cursor in the display, make the setup value effective, return to the previous menu or  
display, and so on.  
Function  
Edit  
Card Selection  
Selects the switch module to be monitored on the LCD. The Card# status indicator  
will change the slot number (card number) when the Card Selection key is pressed.  
This key is not effective when the Card# indicator shows A(auto configuration  
mode).  
Shift  
Exit  
Enables the sub key function. The sub key label has been printed on the front panel  
above the key. The label color is blue.  
Returns the display to the previous menu or display without making the setup  
effective.  
Enter  
Fast  
Makes the setup effective, and returns the display to the previous menu or display.  
Used with the arrow key. Pressing the Fast key and an arrow key speeds up changing  
the setting value.  
arrow keys  
Open/Close  
Moves the cursor on the display, displays another selection for a setup message, or  
changes the value of the digit specified by the cursor.  
For the LED matrix, moves the blink-LED.  
When no blink-LED is on the LED matrix, enters into the LED matrix based switch  
control mode. The blink-LED will appear. In this mode, the arrow keys and Fast key  
are available. Other keys will work to exit this mode.  
When the blink-LED is on the LED matrix, changes the condition, open or close, of  
the switch specified by the blink-LED.  
Agilent B2200 User’s Guide, Edition 2  
3-29  
 
Front Panel Operation  
Setup Menus  
This section explains the following setup menus displayed by pressing the Menu key  
or Selftest key.  
“Setup Menu”  
“Selftest Menu”  
Setup Menu  
Press the Menu key. The setup menu will be displayed on the LCD as shown in the  
following example. The setup menu is used to change the GPIB address, detect the  
couple port automatically, and so on.  
[Hardware Configuration]  
CONFIG SCAN ERROR RMT_DSPL BEEP >  
[Hardware Configuration]  
<SCAN ERROR RMT_DSPL BEEP PEN  
Press the arrow key to move the cursor to the desired function name. Then, press the  
Enter key. The setup message or the sub menu will be displayed. To return to the  
previous menu, press the Exit key.  
3-30  
Agilent B2200 User’s Guide, Edition 2  
   
Front Panel Operation  
Setup Menus  
Function Tree  
The setup menu provides the following functions.  
CONFIG  
ADDRESS  
REVISION Displays the firmware revision.  
UNIT Displays the module information of each slot.  
SCAN Detects and sets the input couple ports automatically.  
ERROR  
DISPLAY  
Sets the GPIB address.  
Displays the error code and error message.  
Clears the error buffer.  
CLEAR  
RMT_DSPL  
LCD  
LED  
Selects on or off for the LCD display in the remote mode.  
Selects on or off for the LED matrix in the remote mode.  
Selects on or off for the beeper.  
BEEP  
PEN  
Enables or disables the light pen.  
Agilent B2200 User’s Guide, Edition 2  
3-31  
Front Panel Operation  
Setup Menus  
CONFIG  
Displays the sub menu that provides the following functions.  
ADDRESS  
Displays the following message.  
GPIB Address = Address  
Press the arrow key to set the desired GPIB address. Then, press the Enter key to  
make the setup effective, or press the Exit key to cancel changing the setup.  
REVISION  
Displays the following message.  
X.XX.XX  
where X.XX.XX shows the firmware revision.  
To return to the previous menu, press the Exit key.  
UNIT  
Displays the following message.  
Slot#: model, X  
where #is 1, 2, 3, or 4, model is the model number of the module, X is the  
revision of the module.  
To display another message, press the arrow key.  
To return to the previous menu, press the Exit key.  
SCAN  
Displays the following message.  
NO  
Press the arrow key to set YES. Then, press the Enter key to detect and set the input  
couple ports automatically, or press the Exit key to cancel this operation. The input  
couple ports will be the input ports the Kelvin cable is connected.  
3-32  
Agilent B2200 User’s Guide, Edition 2  
 
Front Panel Operation  
Setup Menus  
ERROR  
Displays the sub menu that provides the following functions.  
DISPLAY  
Displays the error code and error message, or No Error.  
To return to the previous menu, press the Exit key.  
CLEAR  
Displays the following message. Press the arrow key to set YES. Then, press the  
Enter key to clear the error buffer, or press the Exit key to cancel this operation.  
NO  
RMT_DSPL  
Displays the sub menu that provides the following functions.  
LCD  
Displays the following message. Press the arrow key to set ON (displays all  
status information in the GPIB remote mode) or OFF (displays Remote, Lock,  
and Error only). Then, press the Enter key to make the setup effective, or press  
the Exit key to cancel changing the setup.  
OFF or ON  
LED  
Displays the following message. Press the arrow key to set ON (enables LED  
matrix in the GPIB remote mode) or OFF (disables it). Then, press the Enter key  
to make the setup effective, or press the Exit key to cancel changing the setup.  
ON or OFF  
BEEP  
PEN  
Displays the following message. Press the arrow key to set the beeper ON or OFF.  
Then, press the Enter key to make the setup effective, or press the Exit key to cancel  
changing the setup.  
ON or OFF  
Displays the following message. Press the arrow key to enable or disable the light  
pen. Then, press the Enter key to make the setup effective, or press the Exit key to  
cancel changing the setup.  
DISABLE or ENABLE  
Agilent B2200 User’s Guide, Edition 2  
3-33  
 
Front Panel Operation  
Setup Menus  
Selftest Menu  
Press the Shift key and the Menu key. The selftest menu is displayed. The menu is  
used to execute the selftest and diagnostics.  
[Execute Diagnostics]  
SELF_TEST RELAY_TEST KEY BEEPER >  
[Execute Diagnostics]  
<KEY BEEPER LED PEN GPIB  
Press the arrow key to move the cursor to the desired function name. Then, press the  
Enter key. The setup message or the sub menu will be displayed. To return to the  
previous menu, press the Exit key.  
Function Tree  
This menu provides the following functions.  
SELFTEST  
RELAY_TEST  
KEY  
Executes the controller test or displays test result.  
Executes the relay test or displays test result.  
Executes the front panel interface test or displays test result.  
Executes the beeper test or displays test result.  
Executes the LED matrix test or displays test result.  
Executes the light pen test or displays test result.  
Executes the GPIB test or displays test result.  
BEEPER  
LED  
PEN  
GPIB  
3-34  
Agilent B2200 User’s Guide, Edition 2  
 
Front Panel Operation  
Agilent B2200 User’s Guide, Edition 2  
3-35  
 
Front Panel Operation  
Setup Menus  
KEY  
Displays the sub menu that provides the following functions.  
EXECUTE  
Displays the following message.  
DIAG:KEY Press any key.  
Press any front panel key and confirm the LCD display. For example, if you  
press the Port Function key, the following message must be displayed.  
DIAG:KEY [Port Function]  
Continue this for all keys. After that, press the Enter key twice for the normal  
operation. The test result will be PASS.  
If you find any wrong response, press the Exit key twice. The test result will be  
FAIL.  
Finally, press the Exit key to return to the previous menu.  
RESULT  
Displays the test result. Press the Exit key to return to the previous menu.  
BEEPER  
Displays the sub menu that provides the following functions.  
EXECUTE  
Starts the beeper test, and displays the following message.  
DIAG:BEEP Is beeper making 2 sounds?  
Press the Enter key if the beeper works correctly. The test result will be recorded  
as PASS.  
If you find any problem, press the Exit key. The test result will be recorded as  
FAIL.  
Finally, press the Exit key to return to the previous menu.  
RESULT  
Displays the test result. Press the Exit key to return to the previous menu.  
3-36  
Agilent B2200 User’s Guide, Edition 2  
Front Panel Operation  
Setup Menus  
LED  
Displays the sub menu that provides the following functions.  
EXECUTE  
Starts the LED matrix test, and displays the following message.  
DIAG:LED All LED in Orange?  
Press the Enter key if the all LEDs are lighted in orange. The test result will be  
recorded as PASS.  
If you find any problem, press the Exit key. The test result will be recorded as  
FAIL.  
Finally, press the Exit key to return to the previous menu.  
RESULT  
Displays the test result. Press the Exit key to return to the previous menu.  
PEN  
Displays the sub menu that provides the following functions.  
Before starting the light pen test, connect the light pen to the Agilent B2200.  
EXECUTE  
Starts the light pen test, and displays the following message.  
DIAG:PEN Point any cross point  
Point the light pen to the point you desire and confirm the LCD display. For  
example, if you point the input 5-to-output 10 cross point of the Agilent B2200  
in the auto configuration mode, the following message must be displayed.  
SLOT: 0 INPUT: 5 OUTPUT:10  
Continue this for all cross points. After that, press the Enter key for the normal  
operation. The test result will be recorded as PASS.  
If you find any wrong response, press the Exit key. The test result will be  
recorded as FAIL.  
Finally, press the Exit key to return to the previous menu.  
RESULT  
Displays the test result. Press the Exit key to return to the previous menu.  
Agilent B2200 User’s Guide, Edition 2  
3-37  
Front Panel Operation  
Setup Menus  
GPIB  
Displays the sub menu that provides the following functions.  
EXECUTE  
Displays the following message.  
Open GPIB, then press [Enter]  
Disconnect the cable from the GPIB connector on the rear panel, and open it.  
Press the Enter key to start the GPIB test, and wait until PASS or FAIL is  
displayed. Then press the Exit key to return to the previous menu.  
RESULT  
Displays the test result. Press the Exit key to return to the previous menu.  
3-38  
Agilent B2200 User’s Guide, Edition 2  
4
Programming  
 
Programming  
This chapter describes the automatic control programming of the Agilent B2200.  
“Programming Basics”  
Explains the commands for the fundamental switch control of the Agilent  
B2200.  
“Programming Examples”  
Provides some examples of the programming.  
“Capacitance Compensation”  
Explains how to use the capacitance compensation routine. The routine is a  
function of the Agilent B2200 VXIplug&play driver.  
For the details about SCPI commands, see Chapter 5.  
For the details about VXIplug&play driver, see Chapter 6.  
4-2  
Agilent B2200 User’s Guide, Edition 2  

Mitel RUMOR LG260 User Manual
Melitta Coffeemaker GV09 User Manual
Kenwood KDC MP225 User Manual
HTC TROPHY 7 User Manual
Grindmaster 2450Q User Manual
GBC 5570X User Manual
Fujitsu Computer Drive M9000 User Manual
Freecom Technologies Mobile Hard Drives User Manual
Echo Bear Cat SC3342 User Manual
ACTiSYS ACT IR3200M User Manual