LF414 /LF610
LF414 /LF612
1/2" to 8" (15 to 200 mm)
Field Intelligent Device – Mount-Anywhere Series - Wafer
Electromagnetic Flowmeter
Introduction
The electromagnetic flowmeter uses Faraday’s Law of
electromagnetic induction to measure the process flow.
The device consists of two units: a detector, through
which the fluid to be measured flows and in which
low-level signals proportional to flow rates are
obtained; and a converter, which supplies excitation
current to the detector, and amplifies the signals from
the detector and then processes and converts the signals
into the 4–20mAdc current signal or communication
signal. With the unique patented Mount- Anywhere
magnetic field distribution technology, the meter is
highly immune to upstream flow disturbances.
Combined with a multi-functional converter LF610
(combined type) or LF612 (separate type) equipped
with its patented Noise-Sentry original noise-
suppression circuit and advanced algorithms. The
LF410 has a very high tolerance to noise, giving the
unit a very stable output even for slurry fluid
measurement. IR (Infrared) switches enable parameter
setting of the converter without removing the cover.
Flow direction can be set in either way, and its unique
128 x 128 dot matrix LCD display allows the LCD to
be rotated electronically to 90, 180 and 270 degrees
without opening the cover.
LF410/LF610
LF414/LF610F
LF410
LF414
LF612
LF612F
Figure2. LF410 Mount-Anywhere series
Flowmeters
Certification number
Z01207
Specifications
The AF900 hand-held terminal (HART*1
ꢀOverall Specifications
communicator) can be used to communicate with the
flowmeter from a remote place. PROFIBUS-PA*2
interface is available as an option.
Measurement range in terms of flow velocity:
0 –1.0 ft/s to 0 – 32.8 ft/s (0 – 0.3 m/s to 0 –10 m/s).
*1: HART protocol (Highway Addressable Remote
Transducer) is a communication protocol for industrial
sensors recommended by the HCF (HART
Communication Foundation).
*2: PROFIBUS is the communication protocol for factory
and process automation that the PROFIBUS
Organization recommends. Instead of analog control
with a conventional analog signal (4-20mA), it is
fieldbus which digitizes all signals. Flowmeters support
PROFIBUS-PA.
0 – 0.3 ft/s to 0 – 1.0 ft/s (0 – 0.1 m/s to 0 – 0.3 m/s)
range is available optionally.
Accuracy: ±0.2 % of Rate*
* This pulse output error result is established under standard
operating conditions at Toshiba's flow calibration facility,
Fuchu Japan. (NIST Traceable).
* Individual meter measurement error may vary up to ±0.5% of
Rate at 1.64ft/s (0.5m/s) or more and ±0.3% of rate ±0.039
inch/s (1mm/s) at 1.64 ft/s (0.5 m/s) or less.
Converter
Converter
* Current output: plus ± 8µA (0.05% of span.)
I/O
* Refer to individual calibration data for each individual meter's
measurement error.
Power
supply
Signal cable
Terminal box
Fluid conductivity: 5µS/cm minimum
Fluid temperature:
Power
supply
I/O
14 to 356 °F (–10 to +180°C) : Ceramic type
Note : 248°F (120°C) above is separate type
14 to 248°F (–10 to +120°C) : Teflon PFA
Detector
Detector
Combined type
LF410/LF610
LF414/LF610F
Separate type
LF410/LF612
LF414/LF612F
-
-
Ambient temperature: 4 to 140 °F ( 20 to + 60 °C)
Structure: IP 67 and NEMA 4X Watertight
Figure1. Configuration
TIC-LF414C
Download from Www.Somanuals.com. All Manuals Search And Download.
TIC-LF414C
Note 1: Two outputs (DO1 and DO2) are needed for
4-range switching and forward/reverse 2-range
switching.
Note 2: Normal Open (default set) or Normal Close is
selected for alarm outputs when programming.
When power failure occurs, unit will be fault to
Normal Open.
Digital input DI (opt.)
Signal type: 20 to 30Vdc voltage signal
Input resistance: 2.7kΩ
Number of inputs: one point
DI function — One of the following functions can be
assigned to the optional DI signal.
Range switching — Selects either the higher or
lower range in the unidirectional or bidirectional
2-range setting.
Totalizer control — Starts and stops the built-in
totalizer.
Communications output:
• HART (std.) — Digital signal is superimposed on
4–20mAdc current signal as follows:
Conforms to HART protocol
Load resistance: 240 to 750Ω
Load capacitance: 0.25µF maximum
Load inductance: 4mH maximum
Fixed-value outputs —Outputs fixed-values for
current and pulse outputs.
Zero adjustment — Executes zero adjustment
• PROFIBUS (opt.)
(on-stream at zero flow rate).
Protocol: PROFIBUS-PA
Baud rate: 31.25kbps
Output signals
Current output:
Bus voltage: 9-30VDC
4–20mAdc (load resistance 0 to 750Ω)
Note: The current output cannot be used with the
PROFIBUS-PA communication.
Consumption electric current of bus: less than 16mA
Manufacture Ident-No.: 093BHEX
Standard Ident-No.: 9740HEX
Slave address: 0-126 (Default address is 126)
Profile: Profile Ver.3.01 for Process Control
Devices
Digital outputs — One point (std.) and one more
point is optionally available as follows.
Digital output DO1 (std.):
Output type: Transistor open collector
Number of outputs: One point
Output capacity: 30Vdc, 200mA maximum
Function blocks: AI(Flow)×1 , Totalizer×1
LCD display:
Full dot-matrix 128×128 dot LCD display
(back–light provided)
The data on the LCD inside the converter can rotate
to 90, 180, and 270 degrees by a software, without
rotating the indicator itself. (Combined type only)
Digital output DO2 (opt.):
Output type:
Solid-state relay output (non polarity)
Number of outputs: One point
Output capacity: 150Vdc, 150mA maximum
or 150Vac (peak to peak), 100mA maximum
Parameter settings — Parameters can be set as
follows:
DO1 and DO2 functions — One of the following
functions can be assigned to DO1 (std.) and/or
DO2 (opt.)
• IR Switches: Three key switches are provided to
set configuration parameters.
• Digital communication: The AF900 hand-held
terminal or PROFIBUS is needed to set
parameters.
• Zero adjustment: Zero point adjustment can be
started by pressing the switch in the converter.
• Damping: 0.5 to 60 seconds (selectable in one
• Pulse output (available only for DO1, DO2)
Pulse rate: 3.6 to 36,000,000 pulses/hr (DO1)
3.6 to 360,000 pulses/hr (DO2)
(Over 3,600,000 pulses/hr, auto-setting)
Pulse width: 0.5 to 500ms (but less than half of
the period for 100% flow rate)
Note: The same and simultaneous pulse is not
available between DO1 and DO2.)
second increments)
“Field re-verification” Mag-Prover – Toshiba’s
Zero span calibration tool allows unit to be
re-calibrated and verified using an internal
software program (For more information contact
Toshiba International Corp.)
• Multi-range selection outputs (Note 1)
• High, High high, Low, and/or Low low alarm
outputs (Note 2)
• Empty pipe alarm output
• Digital Output Active Status (DO1 and DO2)
(Note 2)
• Preset count output
• Converter failure alarm output
3
Download from Www.Somanuals.com. All Manuals Search And Download.
TIC-LF414C
Conditions when power fails:
Dimensions and Weights:
See Figure 9 (for Separate type)
Parameter setting values are stored in non–volatile
memory and the values will be restored when the
power returns to normal condition. The outputs and
display will remain as follows when power fails.
MTBF:
Converter: 220,000 hours (25 years) at 77 °F (25 °C)
based on strict military specification
MIL-HDBK-217F
• Current output: 0mAdc
• Digital output:
• LCD display:
• PROFIBUS:
OFF
Detector: 350,000 hours (40 years) at 77 °F (25 °C)
based on strict military specification
MIL-HDBK-217F
No display
No communication
Power supply:
ꢀPED matrix in each flange connection.
The following sizes fall under the category for PED
in each flange connection when the meter ships to
EU. All of them had complied with it from a notified
body.
One of the following can be selected:
• 100 to 240Vac, 50/60Hz (std.)
(allowable voltage 80 to 264Vac)
• 24Vdc (allowable voltage 18 to 36Vdc)
• 110Vdc (allowable voltage 90 to 130Vdc)
Table 1. PED matrix in each flange connection
Flange standard
Meter size
6 to 16 inch
(150 to 400mm)
10 to 16 inch
(250 to 400mm)
6 to 16 inch
Surge protection:
DIN PN 16 and BS 16
Arresters are installed in the power supply and a
current signal output circuit to help protect the meter
from lightning and improve personnel safety.
DIN PN 10 and BS 10
ANSI 150 and JIS10K
(150 to 400mm)
Case: Aluminum alloy (equal to IP67)
Coating: Acrylic resin-baked coating, pearl–gray
colored
Cable connection port:
Cable glands —
LF610 and LF612 without cFMus Approval:
Provided as standard, OD of cableφ11~13mm
Material Nylon 66
G (PF) 1/2 male screws.
LF610F and LF612F with cFMus Approval:
Not provided, 1/2–14NPT male screws are
required.
Applicable diameter —
0.433 to 0.512 inch (11 to 13mm)
Note: When PROFIBUS option is specified, cable gland size is
φ6~8mm for signal cable, φ11~13mm for power cable
Vibration resistance:
No resonance to the following levels of vibration:
• 10 to 150Hz with acceleration of 9.8m/s2
• Vibration of 30Hz with 29.4 m/s2 in 4h in each
direction will not cause any defect to unit.
Note: Avoid using the flowmeter in an environment with
constant vibration.
4
Download from Www.Somanuals.com. All Manuals Search And Download.
TIC-LF414C
Installation
ꢀDimensions
Unit : inch( mm)
Unit : inch( mm)
1.42(36)
1.42(36)
8.86(225)
5.16(131)
5.63(143)
8.86(225)
5.16(131)
5.63(143)
2.52(64)
L1
3.43(87)
L1
ANSI class 150 and class 300 dimensions:
ANSI class 150 and class 300 dimensions:
Meter size
(inch)
L1
L2
D1
Weight
(lb)
Meter size
(inch)
L1
L2
D1
Weight
(lb)
(inch)
(inch)
(inch)
(inch)
(inch)
(inch)
1
3.15
9.48
2.60
approx. 11
1/2
2.76
9.92
1.93
approx. 11
BS16, DIN PN16 and JIS 10K dimensions:
BS16, DIN PN16 and JIS 10K dimensions:
Meter size
(mm)
L1
L2
D1
Weight
(kg)
approx. 5
Meter size
(mm)
L1
L2
D1
Weight
(kg)
approx. 5
(mm)
(mm)
(mm)
(mm)
(mm)
(mm)
25
80
241
66
15
70
252
49
Note: 1 inch = 25.4mm
Note: 1 inch = 25.4mm
Figure 4. LF410/LF610 and LF414/LF610F
flowmeters
Figure 3. LF410/LF610 and LF414/LF610F
flowmeters
Meter size 1"(25mm)
Meter size 1/2"(15mm)
The dimension of L1 is changed when the material of
grounding ring is chosen Pt-Ir or Ta.
Meter size
L1
1/2”(15mm)
1”(25mm)
1 1/2”(40mm)
2”(50mm)
3.03 inch(77mm)
3.74 inch(95mm)
4.53 inch(115mm)
4.96 inch(126mm)
4.96 inch(126mm)
5.35 inch(136mm)
9.53 inch(242mm)
12.28 inch(312mm)
3”(80mm)
4”(100mm)
6”(150mm)
8”(200mm)
5
Download from Www.Somanuals.com. All Manuals Search And Download.
TIC-LF414C
ANSI class 150 and class 300 dimensions
Meter size
(inch)
L1
L2
D1
Weight
(lbs)
(inch)
(inch)
(inch)
Unit : inch(mm)
1-1/2
3.94
4.33
4.33*
4.72*
9.06
10.39
11.02
12.04
13.30
16.02
18.03
3.35
4.02
5.00
6.26
8.50
10.51
approx. 13
approx. 15
approx. 18
approx. 22
approx. 49
approx. 70
2
3
4
6
8
5.63(143)
8.86(225)
5.16(131)
1.42(36)
11.81
BS16, DIN PN16 and JIS 10K dimensions
Meter size
(mm)
L1
L2
D1
Weight
(kg)
(mm)
(mm)
(mm)
40
50
80
100
150
200
100
110
110*
120*
230
264
280
306
338
407
458
85
approx. 6
approx. 7
approx. 8
approx. 10
approx. 22
approx. 32
102
127
159
216
267
300
*When Teflon PFA lining is selected, L1 in the table above
becomes as follows:
L1
ꢁ
L1: 4.53 inch(115mm) for meter size 3"(80mm)
ꢁ
L1: 4.96 inch(126mm) for meter size 4"(100mm)
Note1: Eye bolts are provided at the top for flowmeters
The dimension of L1 is changed when the material of
grounding ring is chosen Pt-Ir or Ta
sized 8"(200mm).
Note2: 1 inch = 25.4mm
Meter size
1/2”(15mm)
1”(25mm)
1 1/2”(40mm)
2”(50mm)
L1
3.03 inch(77mm)
3.74 inch(95mm)
4.53 inch(115mm)
4.96 inch(126mm)
4.96 inch(126mm)
5.35 inch(136mm)
9.53 inch(242mm)
12.28 inch(312mm)
3”(80mm)
4”(100mm)
6”(150mm)
8”(200mm)
Figure 5. LF410/LF610 and LF414/LF610F flowmeters
Meter sizes 1 1/2"(40mm) to 8"(200mm)
6
Download from Www.Somanuals.com. All Manuals Search And Download.
TIC-LF414C
ANSI class 150 and class 300 dimensions:
Unit : inch(mm)
Meter size
(inch)
L1
L2
D1
Weight
(lb)
approx. 7
R (PT)1/2Male Screw (Two Places)
4.96(126)
1.38(35)
(inch)
(inch)
(inch)
3.46(88)
1.57(40)
1/2
2.76
7.05
1.93
3.58(91)
BS16, DIN PN16 and JIS 10K dimensions:
Meter size
(mm)
L1
L2
D1
Weight
(kg)
(mm)
(mm)
(mm)
15
70
179
49
approx. 3
The dimension of L1 is changed when the material of
grounding ring is chosen Pt-Ir or Ta
Meter size
1/2”(15mm)
1”(25mm)
1 1/2”(40mm)
2”(50mm)
L1
3.03 inch(77mm)
3.74 inch(95mm)
4.53 inch(15mm)
4.96 inch(126mm)
4.96 inch(126mm)
5.35 inch(136mm)
9.53 inch(242mm)
12.28 inch(312mm)
2.52(64)
L1
3”(80mm)
Note1: Cable glands are not provided for LF414 of FM and
CSA approved type.
4”(100mm)
6”(150mm)
8”(200mm)
Refer to the part of Cable connection port at
detector.
Note2: 1 inch = 25.4mm
Figure 6. Separate type detectors LF410/LF414
Meter sizes 1/2" (15mm)
7
Download from Www.Somanuals.com. All Manuals Search And Download.
TIC-LF414C
ANSI class 150 and class 300 dimensions:
Unit : inch(mm)
Meter size
(inch)
L1
L2
D1
Weight
(lb)
approx. 7
R(PT)½ Male Screw (Two places)
4.96(126)
1.38(35)
(inch)
(inch)
(inch)
3.46(88)
1.57(40)
1
3.15
6.61
2.60
3.58(91)
BS16, DIN PN16 and JIS 10K dimensions:
Meter size
(mm)
L1
L2
D1
Weight
(kg)
(mm)
(mm)
(mm)
25
80
168
66
approx. 3
The dimension of L1 is changed when the material of
grounding ring is chosen Pt-Ir or Ta
Meter size
1/2”(15mm)
1”(25mm)
1 1/2”(40mm)
2”(50mm)
L1
3.43(87)
L1
3.03 inch(77mm)
3.74 inch(95mm)
4.53 inch(115mm)
4.96 inch(126mm)
4.96 inch(126mm)
5.35 inch(136mm)
9.53 inch(242mm)
12.28 inch(312mm)
Note1: Cable glands are not provided for LF414 of FM and
CSA approved type.
Refer to the part Cable connection port at detector.
Note2: 1 inch = 25.4mm
3”(80mm)
4”(100mm)
6”(150mm)
8”(200mm)
Figure 7. Separate type detectors LF410/LF414
Meter sizes 1" (25mm)
8
Download from Www.Somanuals.com. All Manuals Search And Download.
TIC-LF414C
ANSI class 150 and class 300 dimensions:
Meter size
(inch)
L1
L2
D1
Weight
(lbs)
(inch)
(inch)
(inch)
Unit : inch(mm)
1-1/2
3.94
4.33
4.33*
4.72*
9.06
7.48
8.15
9.13
10.39
12.76
15.16
3.35
4.02
5.00
6.26
8.50
10.51
approx. 9
approx. 11
approx. 13
approx. 20
approx. 47
approx. 78
2
3
4
6
8
R(PT)½Male Screw (Two places)
4.96(126)
1. 38( 35)
3.46(88)
1.57(40)
3.58(91)
11.81
BS16, DIN PN16 and JIS 10K dimensions:
Meter size
(mm)
L1
L2
D1
Weight
(kg)
(mm)
(mm)
(mm)
40
50
80
100
150
200
100
110
110*
120*
230
190
207
232
264
324
385
85
approx. 4
approx. 5
approx. 6
approx. 9
approx. 21
approx. 35
102
127
159
216
267
300
L1
*When Teflon PFA lining is selected, L1 in the table
above becomes as follows:
ꢂL1: 4.53 inch(115mm) for meter size 3"(80mm)
ꢂL1: 4.96 inch(126mm) for meter size 4"(100mm)
Note1: Cable glands are not provided for LF414 of FM and
CSA approved type.
Refer to the part of Cable connection port at
detector.
Note2: Eye bolts are provided at the top for flowmeters
sized 8"(200mm).
Note3: 1 inch = 25.4mm
The dimension of L1 is changed when the material of
grounding ring is chosen Pt-Ir or Ta
Meter size
1/2”(15mm)
1”(25mm)
1 1/2”(40mm)
2”(50mm)
L1
3.03 inch(77mm)
3.74 inch(95mm)
4.53 inch(115mm)
4.96 inch(126mm)
4.96 inch(126mm)
5.35 inch(136mm)
9.53 inch(242mm)
12.28 inch(312mm)
3”(80mm)
4”(100mm)
6”(150mm)
8”(200mm)
Figure 8. Separate type detectors LF410/LF414
Meter sizes 1 1/2"(40mm) to 8"(200mm)
9
Download from Www.Somanuals.com. All Manuals Search And Download.
TIC-LF414C
4 –φ0.43(φ 11)
Unit : inch(mm)
3.23(82)
82
5.94(151)
0.98(25)
25
0.51(13)
13
Attachment
Weight: Approx. 7 lb (3.5 kg)
2.91(74)
LCD display
Excitation cable ground
I/Ocable ground
Gr ound terminal
(M4)
Power supply cable ground
Signa l cable ground
Blind screw
IR Switch
Note: Cable glands are not provided for LF612F cFMus approved type.
Cable ground
Refer to the part Cable connection port at detector.
Note: 1 inch = 25.4 mm
Figure 9. Separate type converter LF612 and LF612F
ꢀExternal Connections
ꢃCombined type LF410/LF610 and LF414/LF610F flowmeters
Instrument Panel : Ordered separately
I/O Cable (CVV-S)
Ground terminal
Current output
(4~20mAdc) or
PROFIBUS
Intermediate
voltage (IV) wire,
5.5mm2 or more.
Digital Output 1
Signal common
for DI and DO
To be grounded
with 100Ω or less
ground resistance.
Digital Output 2
Digital Input
(20 to 30Vdc)
Grounded with
100Ω or less
ground resistance
Grounded with
100Ω or less
ground resistance
Power supply
Power Cable (CVV)
1
*
Power Switch
(External double-pole power switch)
*1 Locate an external double-pole power switch on the power line near the flowmeter within easy reach of operation.
Use the appropriate switch rating as shown below:
Switch rating: 250Vac, 6A or more
In rush current: 15A or more
Figure 10. Combined type LF410/LF610 and LF414/LF610F flowmeters Wiring Diagram
10
Download from Www.Somanuals.com. All Manuals Search And Download.
TIC-LF414C
・Separate type LF410/LF612 and LF414/LF612F flowmeters
Terminal board
IV wire
5.5 mm2 or more
Thick walled steel conduit
Grounded with 100Ω or less
ground resistance
Power switch
(External double-pole power switch )
[Instrument panel : ordered separately]
Grounded with 100Ω or less
ground resistance
Signal cable
(2-wire shielded hard-rubber
sheathed cable)
Power supply
Current output (4~20mAdc)
or PROFIBUS
Digital output 1
Signal common for DI and DO
Connected detector
Digital output 2
Excitation cable
Digital input
(20~30Vdc)
(3-wire hard-rubber cable)
Power cable
(CVV)
Digital input cable
(CVV-S)
Grounded with 100Ω or less
ground resistance
Figure 11. Separate type LF410/LF612 and LF414/LF612F flowmeters wiring Diagram
Table 2. LF610, LF610F, LF612 and LF612F Converters Signal Table
Symbol
L1 (+)
L2 (-)
GND
FG
Description
Cable
Power supply
Power cable (CVV)
Ground (for arrester)
Frame ground
DI
DO1
Digital Input (20~30Vdc)
Digital Output 1
DO2
Digital Output 2
I/O cable (CVV-S)
COM
Signal Common for DI, DO1, DO2
+
-
Current Output (4~20mAdc)
or PROFIBUS
Shielded cable for
PROFIBUS-PA
X
Y
E
A
B
G
Excitation cable
(for LF612, LF612F only)
Excitation Output
Signal Input
Signal cable
(for LF612, LF612F only)
11
Download from Www.Somanuals.com. All Manuals Search And Download.
TIC-LF414C
(7) Only one PROFIBUS-PA cable goes through a
cable gland of the Electromagnetic Flowmeter.
Please use the junction box at system
configuration.
ꢀWiring Precautions
(1) Explosion proof type flowmeters are not provided
cable glands.
Refer to the part Cable connection port at detector
and converter.
300
200
100
(2) Connect the grounding wire (IV wire 5.5mm² or
more) to a good earth ground (100Ω or less ground
resistance). Make the wire as short as possible. Do
not use a common ground shared with other
equipment where earth current may flow. An
independent earth ground is recommended.
(3) The allowable cable lengths between the detector
and converter for the separate type flowmeter
depend on the electrical conductivity of the object
fluid. See Figure 12.
50
30
10
5
3
3
5
10 20 30 50100200
Conductivity (μS/cm)
(4) DO1, DO2 (opt.), and DI (opt.) use the same
common terminal (COM). This COM can not
connect to other equipments which have their own
ground terminal.
Figure 12. Electrical Conductivity and
Cable Length
(Power supply for connecting to DI or DO, etc…)
Need to wire separately.
ꢀ Wiring Precautions (PROFIBUS)
(1) For wiring path, avoid places near electrical
equipment that may cause electromagnetic
induction or electrostatic induction interference
(such as a motor, transformer and wireless
transmitter).
(2) Use a PROFIBUS-PA cable for signal cable. In
addition, make sure to use a shielded cable to
improve noise resistance. Furthermore, installation
of signal cable in metal conduit is recommended.
(3) General PROFIBUS-PA cables are designed for
indoor use where cables are not exposed to
humidity, rain, etc. When you install cables, make
sure to check the operating conditions such as the
operating temperature range of the cable by
contacting its manufacturer.
(4) When you carry out cable end treatment of
PROFIBUS-PA cable, use a dedicated cable
stripper etc. so that the core wire of the cable will
not be nicked or damaged. In addition, for cables,
be careful of allowable maximum bend diameter
etc. (Basically, do not install cables in a way cables
are twisted or bent.).
(5) Consider installing a PROFIBUS-PA arrester in
the communication path of PROFBUS-PA so that
the electromagnetic flowmeter will not be affected
by lightning etc.
(6) The electromagnetic flowmeter is not equipped
with terminating resistors. Use the terminating
resistor unit for PROFIBUS-PA or junction box, if
necessary.
12
Download from Www.Somanuals.com. All Manuals Search And Download.
TIC-LF414C
ꢀMeter Size
ꢀAbout establishment environment
To select the meter size:
Do not store or install the flowmeter :
• Where there is direct sunlight.
• Where excessive vibration or mechanical shock
occurs.
• Where high temperature or high humidity
conditions exist.
See Table 3 to 4 and find meter sizes within the
velocity of 0.3 to 32.8 ft/s (0.1 to 10m/s) for a
specified full-scale (measuring range high limit)
flow. Select one that has its full-scale velocity
between 3.0 and 10 ft/s (1 and 3m/s).
Note: Make sure the full-scale flow rate used for the
final planning stage stays within 32.8 ft/s
(10m/s) in terms of flow velocity.
• Where corrosive atmospheres exist.
• Places that can be submerged under water.
• Where there is a sloped floor. To put the flowmeter
temporarily on the floor, place it carefully with
something, such as a block, to support it so that the
flowmeter will not topple over.
Table 3. Flow rate and flow velocity (English unit)
Unit: gal/min
Size
Flow rate
(inch)
0.328 ft/s 0.98 ft/s 3.0 ft/s
10 ft/s 32.8 ft/s
In areas like the following, there may be the case that
infrared switches do not function correctly. (If these
are unavoidable, use an appropriate cover.)
1/2
1
1 ½
2
3
4
0.2801
0.7781
1.992
3.112
7.967
12.45
28.01
49.80
0.8403
2.334
5.975
9.337
23.90
37.35
84.03
149.4
2.561
7.115
18.21
28.46
72.85
113.8
256.1
455.3
8.532
23.72
60.71
94.86
242.8
379.4
853.8
1,518
28.01
77.81
199.2
311.2
796.7
1,245
2,801
4,980
(1) Where unit (operation panel) is exposed to direct
sunlight, reflection of light onto window pane and
diffused light reflection.
6
8
(2) Where smoke and steam may occur.
(3) Where exposed to direct snow, ice or mud.
Table 4. Flow rate and flow velocity (SI unit)
Unit: m3/h
Flow rate
Size
(mm)
15
25
40
50
80
100
150
200
Ordering Information
0.1 m/s 0.3 m/s 1.0 m/s 3.0 m/s 10 m/s
0.06362
0.1767
0.4523
0.7067
1.809
2.827
6.361
11.31
0.1908
0.5301
1.357
2.120
5.428
8.482
19.08
33.93
0.6361
1.767
4.523
7.067
18.09
28.27
63.61
113.1
1.908
5.301
13.57
21.20
54.28
84.82
190.8
229.3
6.361
17.67
45.23
70.67
180.9
282.7
636.1
1,131
1. When ordering the LF410 series flowmeters, refer
to Tables 6 and 7 (Type Specification Codes).
An entry must be made for each of the columns in
each of these tables.
2. Fluid characteristics:
(1) Type of fluid to be measured and its characteristics
(2) Fluid temperature
(3) Fluid pressure
(4) Electrical conductivity of the fluid
ꢀCalibration Range
3. Measuring range
4. I/O function setting
5. Ordering scope:
If the calibration range is not specified, the standard
range as shown below will be used. If the range is
specified, we will use the specified range for
calibration.
Flow calibration data: (required or not)
6. Other items
Table 5. Standard Flow Range
Specifications other than standard items
Standard flow range
Meter size
Flow rate Flow rate Flow velocity
inch(mm)
Consult a Toshiba representative before
ordering when choosing materials of the wetted
parts such as lining, electrodes, and grounding
rings.
(gal/min)
25
(m³/h)
2
(m/s)
3.144
3.395
3.316
3.537
3.316
3.537
3.144
2.653
1/2 (15)
1 (25)
1 1/2 (40)
2 (50)
75
175
300
650
1,000
2,500
4,500
6
15
25
60
100
200
300
3 (80)
4 (100)
6 (150)
8 (200)
Note: The unit of "gal/min" is not exchanged
(converted) by "m3/h".
13
Download from Www.Somanuals.com. All Manuals Search And Download.
TIC-LF414C
Table 6. Specification Code (Wafer type detector LF410 Series)
Detector category
Normal type
Gr.-A Gr.-B Gr.-C Gr.-D
Model
Specification Code
Description
Ex. type
1
L
L
2
F
F
3
4
4
4
1
1
5
0
4
6
7
8
9
10 11 12 13 14
Normal specification type
√
√
Hazardous location certification type (Note 1)
√
√
Meter size
½"(15mm)
1"(25mm)
1½"(40mm)
2"(50mm)
3"(80mm)
4"(100mm)
6"(150mm)
8"(200mm)
D
E
F
G
H
J
√
√
√
√
√
√
√
√
√
√
√
√
K
L
√
√
√
√
Mounting Style
L
M
P
Detector/Converter combined type (LF410/LF610)
Detector/Converter separate type (LF410/LF612)
Detector/Converter combined type with PED (LF410/LF610) (Note 6)
Detector/Converter separate type with PED (LF410/LF612) (Note 6)
FM and CSA Class - I Division 2 type (Note 1)
Detector/Converter combined type (LF414/LF610F)
Detector/Converter separate type (LF414/LF612F)
√
√
√
√
√
√
√
√
Q
A
B
√
√
√
√
Connection flange standard (Note 2)
C
D
E
F
G
H
J
K
L
Z
ANSI 150
ANSI 300
BS PN 10
BS PN 16
DIN PN 10
DIN PN 16
JIS 10K
JIS 16K
JIS 20K
Other
ꢁ
ꢁ
ꢁ
ꢁ
ꢁ
ꢁ
ꢁ
ꢁ
ꢁ
ꢄ
ꢁ
ꢁ
ꢁ
ꢁ
ꢁ
ꢁ
ꢁ
ꢁ
ꢁ
ꢄ
ꢁ
ꢁ
ꢁ
ꢁ
ꢁ
ꢁ
ꢁ
ꢁ
ꢁ
–
ꢁ
ꢁ
ꢁ
ꢁ
ꢁ
ꢁ
ꢁ
ꢁ
ꢁ
–
Electrode Material (Note 6)
316L stainless steel
Ti (titanium)
Pt-Ir (platinum/iridium)
Ta (tantalum)
B
C
D
E
F
ꢁ
ꢄ
ꢄ
ꢄ
ꢄ
ꢄ
ꢁ
ꢄ
ꢄ
ꢄ
ꢄ
ꢄ
ꢁ
ꢄ
ꢄ
ꢄ
ꢄ
–
ꢄ
ꢄ
ꢄ
ꢄ
ꢁ
–
Hastelloy C (Equivalent)
Other
Z
Lining Material (Note 6)
Alkali-resistant ceramic (Note 3)
Acid-resistant ceramic (Note 3)
Teflon PFA
A
B
C
ꢁ
ꢁ
ꢄ
ꢁ
ꢁ
ꢄ
–
–
ꢁ
–
–
ꢁ
Grounding Ring Material (Note 6)
316 stainless steel
316L stainless steel
C
D
E
ꢁ
ꢄ
ꢄ
ꢁ
ꢄ
ꢄ
ꢁ
ꢄ
ꢄ
ꢁ
ꢄ
ꢄ
Ti (titanium)
F
G
H
Z
Ta (tantalum)
ꢄ
ꢄ
ꢄ
ꢄ
ꢄ
ꢄ
ꢄ
ꢄ
ꢄ
ꢄ
ꢄ
–
ꢄ
ꢄ
ꢄ
–
Pt-Ir (platinum/iridium)
Hastelloy C (Equivalent)
other
Flow and calibration velocity range
1.0 to 32.8 ft/s (standard range calibration)
1.0 to 32.8 ft/s (specified range calibration)
0.3 to 32.8 ft/s (specified range calibration)
Excitation and Signal Cables
not provided
30m cable, provided (Note 4)
other lengths, provided (Note 4)
Coating
A
B
C
ꢁ
ꢄ
ꢄ
ꢁ
ꢄ
ꢄ
ꢁ
ꢄ
ꢄ
ꢁ
ꢄ
ꢄ
A
B
C
ꢁ
ꢄ
ꢄ
ꢁ
ꢄ
ꢄ
ꢁ
ꢄ
ꢄ
ꢁ
ꢄ
ꢄ
A
B
C
D
no coating
ꢁ
ꢄ
ꢄ
ꢄ
ꢁ
ꢄ
ꢄ
ꢄ
–
–
phthalic acid resin coating pearl-gray colored (Note 5)
black tar epoxy resin 0.3 mm
black tar epoxy resin 0.5 mm
ꢁ
ꢄ
ꢄ
ꢁ
ꢄ
ꢄ
Size code explanation: √ :Object ꢁ: Standard ꢄ: Option –: Not available
Note1: Cable glands are not provided. Refer to the part of “Cable connection port” at detector and converter.
Note2: Same shape of flange standard among ANSI 150, ANSI 300, BS10, BS16, DIN10, DIN16, JIS10K, JIS16K, and JIS20K.
Note3: The type of material used is alumina ceramic, this is suitable for both acid and alkali liquids.
This ceramic lining and the electrodes are held with seals, there are two types, one for acids, and the other for alkalis.
Note4: Separate type detector only. Specifying the code “C”, indicate the length of cables from 1 to 300m in 1 meter increments.
Note5: Phthalic acid resin coating is standard for the flowmeter with meter size 1/2”(15mm), 6”(150mm), and 8”(200mm).
Note6: Consult Toshiba before ordering when choose materials at the wetting parts.
Note7: Check the Table 1 whether your chosen meter size meets this directive or not when the meter is shipped to EU. If yes, need to choose this code.
14
Download from Www.Somanuals.com. All Manuals Search And Download.
TIC-LF414C
Table 7. Specification Code for converters
Model
Specification Code
Contents
LF610 LF612
1
L
2
3
6
4
5
6
7
8
9 10 11 12 13 14
type
type
F
1
Electromagnetic flowmeter converter
0
2
Combined (Integral) type
Separate (Remote) type
ꢁ
–
–
ꢁ
Purpose
A
F
Standard
cFMus class I, Division 2 approved
ꢁ
ꢄ
ꢁ
ꢄ
Shape
Integral type with case
ꢁ
A
B
–
ꢁ
–
Separate type with case
Converter mounting fitting
A
C
E
None
ꢁ
–
–
ꢄ
ꢁ
ꢄ
Panel, Accessory for wall mounting (BNP material: SUS304)
Accessory for pipe installation (BNP material: SUS304)
Digital input/output
1
2
Digital output points 1 (DO1)
ꢁ
ꢄ
ꢁ
ꢄ
Digital output points 2 (DO1+DO2) +Digital input point 1 (DI)
Current output and Communication function
Current output + HART communication
PROFIBUS communication (Current output is not usable)
Power supply
1
2
ꢁ
ꢄ
ꢁ
ꢄ
1
2
3
100Vac-240Vac, 50/60Hz
24Vdc
110Vdc
ꢁ
ꢄ
ꢄ
ꢁ
ꢄ
ꢄ
Instruction manual
E
English
ꢁ
ꢁ
Code explanation: ꢁ: Standard ꢄ: Option
–: Not available
15
Download from Www.Somanuals.com. All Manuals Search And Download.
TIC-LF414C
Specifications are subject to change without notice.
Printed in Japan 2008-6 (TDOC)
© TOSHIBA Corporation 2008
All Rights Reserved.