Teledyne Pet Fence M201E User Manual

MANUAL ADDENDUM  
MODEL M201E  
AMMONIA ANALYZER  
(For use with the M200E Operators Manual, P/N 04410)  
TELEDYNE ADVANCED POLLUTION INSTRUMENTATION  
(TELEDYNE-API)  
9480 CARROLL PARK DRIVE  
SAN DIEGO, CA 92121-5201  
TOLL-FREE:  
FAX:  
800-324-5190  
858-657-9816  
TEL:  
858-657-9800  
E-MAIL:  
WEB SITE:  
Copyright 2006-2012  
Teledyne Advanced Pollution Instrumentation  
05206H DCN5910  
07 May 2012  
Download from Www.Somanuals.com. All Manuals Search And Download.  
ABOUT THIS MANUAL  
This manual is an addendum that supplements the M200E NOx Analyzer manual, p/n 04410; it  
includes document(s) as listed below.  
Part No.  
Name/Description  
M201E Spare Parts List (in Section 6.0)  
Note: Please refer to our Website or Sales for more  
recent updates.  
06185H  
NOTE  
We recommend that this manual and the M200E manual be read in their  
entirety before any attempt is made to operate the instrument.  
REVISION HISTORY  
M201E, MANUAL ADDENDUM, PN 05206  
Date  
To Rev  
DCN  
Change Summary  
07 May 2012  
14 May 2010  
H
G
5910 Updated specs and valve port labels  
5731 Administrative additions  
Added thermocouple replacement  
instructions  
Updated catalytic cartridge illustration  
17 Nov 2009  
F
5588  
Updated pneumatic diagrams  
Updated Spare Parts List Rev E to F  
05206H DCN5910  
i
Download from Www.Somanuals.com. All Manuals Search And Download.  
This page intentionally left blank.  
ii  
05206H DCN5910  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model M201E NH3 Analyzer Operator Manual  
M201E Ammonia Analyzer  
SAFETY MESSAGES  
Your safety and the safety of others is very important. We have provided many safety messages  
throughout this manual; please read these messages carefully.  
A safety message alerts you to potential hazards that could hurt you or others. Safety messages  
are associated with a safety alert symbol. These symbols are found in both the M200E and  
M201E manuals and are also located inside the instrument. The definition of these symbols is  
described below:  
GENERAL WARNING/CAUTION: Refer to the instructions for details on the  
specific danger.  
CAUTION: Hot Surface Warning  
WARNING: Electrical Shock Hazard  
Technician Symbol: All operations marked with this symbol are to be performed  
by qualified maintenance personnel only.  
Electrical Ground: This symbol inside the instrument marks the central safety  
grounding point for the instrument.  
CAUTION The analyzer should only be used for the purpose  
and in the manner described in this manual.  
If you use the analyzer in a manner other than that for which  
it was intended, unpredictable behavior could ensue with  
possible hazardous consequences.  
05206H DCN5910  
iii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
This page intentionally left blank.  
iv  
05206H DCN5910  
Download from Www.Somanuals.com. All Manuals Search And Download.  
TABLE OF CONTENTS  
1.1  
1.1.1  
1.1.2  
1.2  
1.3  
1.4  
Theory of Operation...................................................................................................... 10  
Minimizing PMT Drift.......................................................................................... 11  
Purging the Reaction Cell..................................................................................... 12  
Special Considerations For Ammonia Measurement ................................................... 12  
Sample Filtration........................................................................................................... 13  
M201E Analyzer Specifications ................................................................................... 14  
2.1  
2.2  
2.3  
Zeroing The Analyzer................................................................................................... 16  
Spanning The Analyzer With Nitric Oxide Gas ........................................................... 17  
Spanning The Analyzer With Ammonia Gas ............................................................... 20  
3.0 CONFIGURABLE ANALOG OUTPUT OVERVIEW ...........................25  
3.1  
3.2  
3.3  
3.4  
3.5  
Analog Output - Electronic Configuration ................................................................... 27  
Analog Output Signal Type and Range Selection ........................................................ 28  
Selecting Auto Or Manual Analog Output Calibration ................................................ 29  
Manually Calibrating Analog Output Signal Levels .................................................... 32  
Analog Output Recorder Offset Adjustment ................................................................ 34  
4.1  
4.2  
4.3  
4.4  
M501 NH3, Maintenance ............................................................................................. 35  
Replacing the Catalytic Cartridge................................................................................. 36  
Replacing the Thermocouple........................................................................................ 37  
AutoZero Flow Check................................................................................................... 39  
5.1  
5.2  
Alarm ............................................................................................................................ 41  
Caution.......................................................................................................................... 41  
7.0 CALCULATING MOLY CONVERTER EFFICIENCIES ......................47  
7.1  
7.2  
7.3  
7.4  
Purpose:......................................................................................................................... 47  
TOOLS:......................................................................................................................... 47  
PARTS:......................................................................................................................... 47  
PROCEDURE............................................................................................................... 47  
LIST OF FIGURES  
Figure 1-1. M201E Pneumatic Connection Diagram with External Calibrator Option (ECO)...... 8  
Figure 1-2. M201E and M501 NH3, Pneumatic Flow .................................................................... 9  
Figure 1-3. M201E and M501 NH3, with Zero and Span Valve Options Pneumatic Flow.......... 10  
Figure 3-1. Analog Output Connector .......................................................................................... 25  
Figure 3-2. Analog Output Signal Type ....................................................................................... 28  
Figure 3-3. Auto Or Manual Analog Output Calibration ............................................................. 30  
Figure 3-4. Analog Output Group Calibration.............................................................................. 31  
Figure 3-5. Setup For Calibrating Analog Output Signal Levels ................................................. 32  
05206H DCN5910  
v
Download from Www.Somanuals.com. All Manuals Search And Download.  
Figure 3-6. Analog Output Manual Calibration............................................................................ 33  
Figure 3-7. Analog Output Recorder Offset Calibration .............................................................. 34  
Figure 4-1. Catalytic Cartridge ..................................................................................................... 36  
Figure 4-2. Thermocouple Location ............................................................................................. 37  
Figure 5-1. M201E Functional Block Diagram............................................................................ 42  
LIST OF TABLES  
Table 1-1. M201E Operating Specifications ................................................................................ 14  
Table 2-1. Zero Calibration Procedure – Zero Gas Through the SAMPLE Port ......................... 16  
Table 2-2. Zero Calibration Procedure - Zero Gas Through ZERO Port ..................................... 17  
Table 2-3. NO Calibration Procedure – NO Gas Through the SAMPLE Port............................. 18  
Table 2-4. NO Calibration Procedure - NO Gas through the SPAN Port.................................... 19  
Table 2-5. Confirming Ammonia Converter Efficiency ............................................................... 21  
Table 2-6. NH3 Calibration Procedure – NH3 Gas Through The SAMPLE Port........................ 22  
Table 2-7. NH3 Calibration Procedure – NH3 Gas through the SPAN Port................................ 23  
Table 3-1. Analog Output Data Default Settings.......................................................................... 25  
Table 3-2. Analog Output Pin-Outs.............................................................................................. 26  
Table 3-3. Analog Output Voltage Range Min/Max .................................................................... 27  
Table 3-4. Analog Output Current Loop Min/Max ...................................................................... 27  
Table 3-5. ZERO/SPAN Analog Output Signal Calibration Tolerances...................................... 32  
Table 4-1. Preventative Maintenance Schedule............................................................................ 35  
Table 4-2. M501 NH3, Converter Rebuild Parts List.................................................................... 36  
Table 4-3. Analyzer Flow Check.................................................................................................. 39  
vi  
05206H DCN5910  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1.0 M201E AMMONIA ANALYZER  
This manual is to be used in conjunction with the M200E NOx Analyzer  
manual, p/n 04410 supplied with this instrument. It is important that you  
familiarize yourself with the workings of the NOx analyzer before proceeding  
with NH3 measurements.  
The M201E ammonia analyzer consists of a modified M200E nitrogen oxide  
analyzer and an M501 NH3, ammonia converter. The analyzer measures the  
following individual gas concentrations: TNx (NH3 and NOx), NOx (NO and NO2)  
and NO in the sample gas.  
The ammonia concentration is calculated by subtracting the NOX  
reading from the TNx measurement.  
The nitrogen dioxide concentration is calculated by subtracting the NO  
reading from the NOx measurement.  
The instrument is available in three pneumatic configurations:  
The M201E with External Calibrator Option (ECO) is shown in Figure  
1-1.  
The basic system is shown in Figure 1-2.  
The M201E with Zero/Span valves is shown in Figure 1-3.  
Note: For an M201E analyzer with Zero/Span option you must connect the  
umbilical cord P/N 02255 between the analyzer and the external M501 NH3,  
converter assembly.  
If the analyzer is purchased without the Zero/Span option then the zero and span  
calibration gases must be individually applied to the sample port. Remember, the  
gases must be clean/dry and supplied at ambient pressure.  
CAUTION  
Do not pressurize the Sample/Span/Zero port above ambient  
05206H DCN5910  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
M201E Ammonia Analyzer  
Teledyne API Model M201E NH3 Analyzer Operator Manual  
Figure 1-1. M201E Pneumatic Connection Diagram with External Calibrator Option (ECO)  
8
05206H DCN5910  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Teledyne API Model M201E NH3 Analyzer Operator Manual  
M201E Ammonia Analyzer  
Figure 1-2. M201E and M501 NH3, Pneumatic Flow  
05206H DCN5910  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
M201E Ammonia Analyzer  
Teledyne API Model M201E NH3 Analyzer Operator Manual  
Figure 1-3. M201E and M501 NH3, with Zero and Span Valve Options Pneumatic Flow  
1.1 THEORY OF OPERATION  
The Teledyne-API Model M201E analyzer measures ammonia by oxidizing it to  
nitric oxide by the following reaction:  
4NH3 + 5O2 4NO + 6H2O  
The resulting nitric oxide is then measured by the chemiluminescent reaction with  
ozone. Consult Section 10.1 in the M200E manual (provided with this  
instrument) for more details on the nitric oxide measurement.  
The analyzer uses two converters to oxidize the different sample gases. A high  
temperature catalytic converter, the M501 NH3,, converts NH3 and NOx into NO  
creating the TNx channel. A second converter, housed inside the M201E  
analyzer and consisting of heated molybdenum, converts all of the NOx in the  
sample to NO producing the NOx channel. The nitric oxide channel is measured  
while bypassing both the M501 NH3, and molybdenum converter. A Nafion® drier  
operated in reflux mode, is installed prior to the molybdenum converter and the  
AutoZero valve. The drier removes ammonia and water from the switched  
stream. The ammonia would have otherwise been converted by the molybdenum  
10  
05206H DCN5910  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Teledyne API Model M201E NH3 Analyzer Operator Manual  
M201E Ammonia Analyzer  
and registered as NOx. This location provides the drier with continuous flow,  
thereby allowing it to stabilize faster.  
Note: Due to the Nafion® drier, the analyzer has a maximum ammonia range of  
2.0 PPM.  
The calculated gas concentrations, ammonia and nitrogen dioxide, are computed  
from the differences between the measured channels, as outlined in the following  
equations:  
NO (Nitric Oxide) Concentration = Measured  
TNx (Total Nitrogen) Concentration = Measured (NH3 + N02 + NO)  
NOx (Total Nitrogen – Ammonia) Concentration = Measured (NO2 + NO)  
NH3 (Ammonia) Concentration = Calculated as follows:  
[(TNx - NOx) / (NH3_CE_FACTOR1)]  
NO2 (Nitrogen Dioxide) Concentration = Calculated as follows:  
[(NOx – NO) / (NO2_CE_FACTOR1)]  
Note: The nitrogen dioxide efficiency factor (NO2_CE_FACTOR1) must be  
calculated by conducting a GPT (gas phase titration) as described in Section  
8.3.3 of the M200E manual and Section 7.0 of this manual. This value gives a  
good indication of how well the molybdenum is converting nitrogen dioxide into  
nitric oxide. In a properly functioning analyzer the value should be close to 1.0. It  
is recommended that the molybdenum efficiency be checked every three months.  
The ammonia converter efficiency factor (NH3_CE_FACTOR1) is discussed in  
more detail in Section 2.3 of this manual. The conversion efficiency of the M501  
NH3, should be checked prior to starting long term tests. Both efficiency factors  
can be accessed through the analyzer VARS MENU.  
The actual formula for computation of the gas concentrations is more  
complicated than the above equations, as it includes sample, reaction cell  
pressure changes and averaging the PMT signal. Then the zero offset and  
slopes are applied to the separate channels (TNx, NOx, NO) to determine the  
concentrations. Concentration compensation occurs while the variable  
TP_FACTOR under the VARS MENU is switched ON. Otherwise the displayed  
concentration is uncompensated. It is recommended that the variable  
TP_FACTOR remain on at all times.  
1.1.1  
Minimizing PMT Drift  
In order to account for PMT drift in the analyzer, the AutoZero valve switches  
once a minute allowing the analyzer to read zero background. The AutoZero  
valve directs the sample gas stream to completely bypass the reaction cell, while  
simultaneously filling the reaction cell with Ozone for dark noise measurement.  
This is then subtracted as a measurement offset from the raw PMT concentration  
05206H DCN5910  
11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
M201E Ammonia Analyzer  
Teledyne API Model M201E NH3 Analyzer Operator Manual  
signal. This process improves zero baseline stability by minimizing the effect of  
PMT sensor drift.  
1.1.2  
Purging the Reaction Cell  
As with many chemical reactions the conversion of ammonia in the presence of  
other oxides of nitrogen is complicated. It is important to note that the valve  
DWELL time for an AZERO measurement has a default setting of 8 seconds.  
Shortening this value may not allow enough time to properly purge the reaction cell  
of excess nitric oxide from the previous measurements.  
In the molybdenum converter operating at 315oC the following significant reactions  
are taking place:  
Mo + NO2 MoO3 + NO ~100% Efficiency  
The M501 NH3, ammonia converter operates at 825oC. At this high temperature,  
several reactions occur:  
NO NO  
NO2 NO  
NH3 NO  
NH3 NO2  
Loss = ~ 3%  
Efficiency = ~97%  
Efficiency = ~97%  
Efficiency = ~5%  
As can be seen from the above reactions, the calculation of the ammonia  
concentration and overall calibration of the instrument must be done carefully, if  
accurate ammonia concentrations are to be measured.  
1.2 SPECIAL CONSIDERATIONS FOR AMMONIA MEASUREMENT  
Ammonia is a difficult gas to measure due to its chemical characteristics. The  
gas tends to adsorb onto surfaces and diffuse into many materials. The following  
precautions should be observed when designing ammonia sampling systems and  
connecting them to the M201E analyzer:  
Do NOT use copper tubing or fittings designed for household plumbing.  
Use ONLY Chromatography grade (cleaned, passivated) stainless steel  
tubing.  
Use ONLY Glass tubing for sample inlet manifold.  
ALWAYS keep the tubing from the M702SS calibrator to the analyzer as  
short as possible. USE stainless steel tubing throughout, especially from the  
ammonia calibration bottle to the calibrator.  
If possible HEAT the sample line and DRY the sample gas.  
Sample filters Will cause response delays on both the rise and fall of the  
sample signal.  
12  
05206H DCN5910  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Teledyne API Model M201E NH3 Analyzer Operator Manual  
M201E Ammonia Analyzer  
These rules apply also to your calibrator’s internal plumbing. It is highly  
recommended that you use a M702SS calibrator in conjunction with the M201E.  
The calibrator has stainless steel plumbing throughout. The M700 calibrator  
contains Teflon tubing and internal MFC’s that can be affected by the ammonia  
gas, therefore, the M700 calibrator is not recommend for use with the M201E  
analyzer. It is important to keep the sampling system well maintained.  
1.3  
SAMPLE FILTRATION  
The instrument can be provided with an optional stainless steel sample filter P/N  
05571. For minimum response time operation, the instrument should be supplied  
with clean sample gas at ambient pressure.  
An additional filter resides in the M201E pneumatic pathway. The filter is a  
¼” diameter stainless steel sintered filter used to protect the reaction cell flow  
control orifice from plugging. This filter will plug rapidly if contaminated sample  
gas is not filtered before entering the analyzer.  
NOTE: In situations where the sample gas is known to be contaminated with  
particulate material it is advisably to add extra filtration. However, it must be  
noted that the response time is directly extended through filter additions.  
05206H DCN5910  
13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
M201E Ammonia Analyzer  
Teledyne API Model M201E NH3 Analyzer Operator Manual  
1.4 M201E ANALYZER SPECIFICATIONS  
Table 1-1. M201E Operating Specifications  
0-50 to 0-2000 ppb in 1 ppb increments (selectable, independent  
NH3, NO, NO2, NOx ranges supported)  
PPB, PPM, microgram/m3, milligram/m3 (user selectable)  
< 0.5 ppb RMS  
Ranges  
Measurement Units  
Noise at Zero1  
Noise at Span1  
Lower Detectable  
Limit2  
< 1.0% of reading above 50 ppb  
1 ppb RMS  
Zero Drift3, 6  
2 ppb / 24 hours  
< 1.0% FS Range / 24 hours  
40 seconds  
90% 300 seconds  
90% 300 seconds  
Span Drift6  
Lag Time  
Rise Time4  
Fall Time4  
1000 cm3/min + 10% (500 cm3/min bypass to vacuum manifold,  
500 cm3/min to reaction cell)  
NO calibration 1% of full scale  
Sample Flow Rate  
Linearity  
NH3 calibration 2% of full scale  
Temp Range  
15-40oC  
Dimensions HxWxD  
Weight, Analyzer  
Weight, Converter  
Weight, Pump  
7” x 17” x 23.6” (18cm x 43cm x 61cm)  
43 lbs (20 kg)  
24 lbs (11 kg)  
16 lbs (7 kg)  
100V ~50/60 Hz, 120V ~60 Hz, 220V ~50Hz, 240V ~50 Hz, 125  
watts  
230V ~50 Hz, 125 watts  
110V ~60 Hz, 220V ~50 Hz, 240V ~50 Hz, 295 watts  
230 V ~50 Hz, 2.5 A peak  
Installation Category Pollution Degree 2, Over-voltage Category II  
4 user configurable outputs, each can be configured as  
0.1, 1V, 5V or 10V. Three outputs convertible to 4-20mA isolated  
current loop.  
Power, Analyzer  
Power, Analyzer5  
Power, Pump  
Power, Pump CEMark5  
Environmental  
Analog Output  
Analog Output  
Resolution  
1 part in 4096 of selected full-scale voltage (12 bit)  
1x RS-232, 1 x RS-485 or RS-232 or Ethernet(optional).  
Communication speed: 300 – 115200 baud (user selectable)  
8 status outputs from opto-isolators, 7 defined, 1 spare  
Serial I/O  
Status Option  
1 As defined by USEPA  
2 Defined as twice the zero noise level  
3 At constant temperature and voltage  
4 When pneumatics are conditioned with NH3 overnight and the gases (zero air to NH3) are manually  
switched at the sample inlet at the rear of the M501.  
5 Electrical rating for CE Mark Compliance  
6 Applies when sampling NH3; better results expected for NO/NOx gas measurements.  
14  
05206H DCN5910  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
2.0 CALIBRATION PROCEDURE  
This section begins with a high-level overview of the calibration procedure for the  
M201E analyzer. Details are provided starting in Section 2.1:  
First, take the time to read the M200E manual to familiarize yourself with  
the Chemiluminescence process. It is important to remember that the  
analyzer is merely measuring different levels of nitric oxide sample gas on  
three separate channels (TNx, NOx, NO). The ammonia and nitrogen  
dioxide concentrations are then calculated using this information.  
Assemble the M201E analyzer according to the pneumatic configurations  
outlined in Figure 1-1. Remember: DO NOT pressurize the sample, span or  
zero ports (Zero/SPAN valves are optional) during calibration. Allow the  
analyzer to pull the gas through the system using the vacuum pump.  
CAUTION!  
If the presence of ozone is detected at any time, call Teledyne API Customer Service as  
soon as possible:  
800-324-5190 or email: [email protected]  
Next, zero the analyzer using an approved zero air source, such as, bottled  
zero air, nitrogen or zero air generated by a Teledyne-API M701 zero air  
source. This sets the zero offset for the three individual channels  
(TNx_OFFS, NOx_OFFS, NO_OFFS). Confirm that all displayed  
concentrations read zero.  
Span the analyzer using bottled nitric oxide gas diluted to a level set to 80%  
of the expected sample range, preferably using a M702 Calibrator. This  
sets the slope for the three individual channels (TNx_SLOPE, NOx_SLOPE,  
NO_SLOPE). Confirm that displayed concentrations TNx , NOx and NO  
display the correct span concentration.  
If you haven’t done so recently, conduct a GPT (gas phase titration) and  
confirm the molybdenum is functioning accordingly. Section 8.3.3 of the  
M200E manual outlines the GPT procedure. Section 7.0 of this manual  
consists of a service note for checking the molybdenum converter. There  
should be no reason to adjust the molybdenum efficiency factor on a new  
analyzer. If a diluted bottled of nitrogen dioxide gas is used to determine  
the molybdenum efficiency, allow enough time for the span value to  
stabilize. Nitrogen dioxide exhibits similar hold up issues as ammonia gas.  
Therefore, it may take a number of hours before the NOx reading  
stabilizes.  
Span the analyzer using bottled ammonia gas diluted to a level set to 80%  
of the expected sample range, preferably using a M702 Calibrator. Note:  
05206H DCN5910  
15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Calibration Procedure  
Teledyne API Model M201E NH3 Analyzer Operator Manual  
The first time bottled ammonia is connected to the gas dilution system,  
regulator/calibrator, the TNx reading may take a number of hours to  
stabilize.  
2.1 ZEROING THE ANALYZER  
The analyzer can be zeroed by either applying zero air straight to the sample port  
or through the optional Zero/Span valves, if installed. It is important to remember  
that if the analyzer was previously sampling ammonia gas, prior to conducting a  
Zero calibration, it will take some period of time before the zero baseline is  
reached. The process of zeroing the analyzer consists of delivering dry zero air  
to the analyzer. The operator then manually zeros the TNx and NOx channels  
individually through the user interface. The following Tables outline the steps  
necessary to zero the analyzer.  
Table 2-1. Zero Calibration Procedure – Zero Gas Through the SAMPLE Port  
Important: Each channel (TNx and NOx) must be zeroed individually  
Step No.  
Action  
Comment  
1.  
Press CAL  
The M201E enters the calibrate mode from sample  
mode. Confirm zero gas is flowing past the sample  
port.  
2.  
3.  
Channel Selection Press TNX or NOx then Press ENTR.  
Range Selection  
Press LOW or HIGH range then Press ENTR.  
Always calibrate the LOW range, section 3.0 below.  
4.  
5.  
NH3 STB Reading Wait for the displayed stability reading to fall below  
1.0 PPB. The lower the stability reading the better the  
zero value.  
Press ZERO  
If you change your mind after pressing ZERO, you  
can still press EXIT without zeroing the instrument.  
You don’t have to enter concentration values. The  
analyzer knows to apply 0 PPB.  
6.  
Press ENTR  
Pressing ENTR actually adjusts the Offset value for  
the selected channel.  
7.  
8.  
Press EXIT  
The M201E returns to sampling mode.  
Check  
Concentrations  
/Offsets  
All displayed concentrations should read zero. The  
offset values should be close to zero (-20 to 150 mv)  
16  
05206H DCN5910  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Teledyne API Model M201E NH3 Analyzer Operator Manual  
Calibration Procedure  
Table 2-2. Zero Calibration Procedure - Zero Gas Through ZERO Port  
Step No.  
Action  
Comment  
1.  
Press CALZ  
The M201E enters the calibrate mode from sample  
mode. The zero gas is supplied through the ZERO  
inlet port on the rear panel of the M501 NH3,.  
2
Table 2.1  
Follow Steps 2 to 8 in Table 2-1.  
Notes:  
Since the zero gas concentration is defined as 0 ppb, it is not necessary to  
enter the expected concentration values.  
Both the TNx and NOx channels must be zeroed individually. When you  
zero the NOx channel both the offsets for NOx and NO are configured. All  
readings should display zero following this action.  
It is recommended that the calibration of both the TNx and NOx channels  
be done at one time.  
Always confirm that excess zero air is flowing past the sample or zero  
port, if the Zero/Span option is present, on the rear of the M501 NH3,  
converter. There should always be some excess flow. If insufficient flow is  
supplied to the analyzer ambient gas will be entrained and the zero offset  
values will be incorrect. However, it is important not to over pressurize the  
analyzer.  
It is a good idea to allow the analyzer to sample zero gas for extended  
periods following zero calibration. Large drifts can signify leaks or indicate  
an issue with the zero air source.  
2.2 SPANNING THE ANALYZER WITH NITRIC OXIDE GAS  
The analyzer can be spanned by either applying nitric oxide gas straight to the  
Sample port or through the Zero/Span valve option, if installed. It is important to  
remember that if the analyzer was previously sampling ammonia gas, prior to  
conducting a nitric oxide span calibration, it will take some period of time before  
the TNx channel stabilizes. The process of spanning the analyzer consists of  
diluting bottled gas, using a calibrator and delivering to the analyzer. The  
05206H DCN5910  
17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Calibration Procedure  
Teledyne API Model M201E NH3 Analyzer Operator Manual  
operator then manually spans the TNx and NOx channels individually through the  
user interface. The following Tables outline the manual steps necessary to span  
the analyzer.  
Table 2-3. NO Calibration Procedure – NO Gas Through the SAMPLE Port  
Important: Each channel (TNx and NOx) must be Spanned individually  
Step No.  
Action  
Comment  
1.  
Press CAL  
The M201E enters the calibrate mode from sample  
mode. Confirm that excess NO gas is flowing past the  
sample port.  
2.  
3.  
Channel Selection Press TNX or NOx then Press ENTR.  
Range Selection  
Press LOW or HIGH range then Press ENTR.  
Always calibrate the LOW range, Section 3.0below.  
4.  
Press CONC  
If NOx was chosen previously then the following will  
be displayed:  
NOx NO CONV  
Exit  
Press NOx, If you are delivering 450 PPB of NO then  
Enter 450 PPB, Press ENTR.  
Repeat the same steps for NO.  
Press EXIT  
Once stability has been achieved,  
Press SPAN, ENTR, EXIT  
NOTE: Both NOx and NO should read 450 PPB.  
____________________________________________  
If TNx was chosen previously then the following will  
be displayed:  
TNx CONV  
Exit  
Press TNx enter 450 PPB, Press ENTR  
Press EXIT  
Once stability has been achieved,  
Press SPAN, ENTR, EXIT  
TNx should read 450 PPB.  
5.  
Check Slopes  
The slopes should be close to 1.0 +/- 0.300  
18  
05206H DCN5910  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Teledyne API Model M201E NH3 Analyzer Operator Manual  
Calibration Procedure  
Table 2-4. NO Calibration Procedure - NO Gas through the SPAN Port  
Step No.  
Action  
Comment  
1.  
Press CALS  
The M201E enters the calibrate mode from sample  
mode. The NO gas is supplied through the SPAN inlet  
port on the rear panel of the M501 NH3,.  
2
Table 2.3  
Follow Steps 2 to 5 in Table 2-3.  
Notes:  
Both the TNx and NOx channels must be spanned individually with nitric  
oxide gas. When you span the NOx channel both the NO and NOx  
concentrations must be manually inputted.  
It is recommended that the calibration of both the TNx and NOx channels  
be done at the same time.  
Always confirm that SPAN gas is flowing past the sample or span port on  
the rear of the M501 NH3, converter. There should always be some  
excess flow. If insufficient flow is supplied to the analyzer ambient gas will  
be entrained and the span values will be incorrect. However, it is important  
not to over pressurize the analyzer.  
It is a good idea to allow the analyzer to sample span gas for extended  
periods following span calibration. Large drifts can signify leaks or indicate  
an issue with the span gas source.  
If after spanning with nitric oxide, the analyzer slopes are out of range  
conduct the following:  
. Confirm the gas sources are good. Usually the bottle is certified to  
a known level of nitric oxide. The bottle should be balanced with  
nitrogen. Try not to use the same regulator for nitric oxide as was  
used on ammonia. This also holds for the tubing between the  
different bottles and the mixing source (calibrator).  
. Confirm the zero reading is good by delivering Zero gas.  
. Confirm that the concentration delivered to the analyzer is equal to  
what you entered under the CONC menu NOx and TNx.  
. Leak check both the M501 NH3, and analyzer together.  
. Using section 11.6.5 of the M200E manual confirm that the PMT  
sensor hardware calibration is set correctly.  
05206H DCN5910  
19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Calibration Procedure  
Teledyne API Model M201E NH3 Analyzer Operator Manual  
2.3 SPANNING THE ANALYZER WITH AMMONIA GAS  
The most important criterion an individual can avail themselves of when spanning  
an analyzer with ammonia gas is “patience”. Ammonia is a very sticky gas and  
the response of the analyzer/calibration system depends on a number of factors.  
The following lists some important points the user should be aware of when  
calibrating:  
Is this the first time the analyzer/calibration system has seen ammonia gas  
in the last couple of days? If so, spanning to a fixed value will take  
considerably longer than normal: possibly 12 hours to completely stabilize.  
Is the tubing that is delivering span gas from the calibration bottles, made  
of stainless steel? Is the tubing from the rear of the calibrator to the  
analyzer, made of stainless steel? Where possible use stainless steel  
tubing.  
Note: The operator is responsible for delivering a set amount of ammonia  
to the analyzer, calculating the converter efficiency of the M501 NH3, and  
entering the efficiency value through the user interface, if required.  
Is the environment where the analyzer resides undergoing large  
temperature swings (+/- 5 C about the norm Temperature)? If so the  
sample line can absorb and desorb ammonia at concentration rates that  
are distinguishable to the analyzer. If possible, heat the sample line to a  
constant maximum ambient temperature. Otherwise the FILT_SIZE = 42  
Samples variable located under the VARS MENU can be lengthened.  
However, lengthening the filter size slows the rise/fall response time of the  
analyzer.  
Does the sampling environment contain large amounts of particulates? If  
so, it may be necessary to add the sampling filter option. As outlined  
previously, this will affect the analyzer response.  
One method to increase the response of both the sampling system and  
the analyzer is to flow a larger concentration of ammonia gas through the  
pneumatics for a couple of hours. For example, if the expected range is  
500 PPB of ammonia, then flowing 1.0 PPM speeds up the absorption  
which occurs throughout the pneumatics. The long delays in ammonia  
response occur when the analyzer hasn’t sampled the gas for some time.  
Ammonia desorbs from the pneumatics, these surfaces need to be  
rewetted before span stability is achieved.  
Prior to conducting the ammonia calibration, confirm that the ammonia  
converter efficiency is initially set to 1.000 using the following menu key  
selections:  
20  
05206H DCN5910  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Teledyne API Model M201E NH3 Analyzer Operator Manual  
Calibration Procedure  
Table 2-5. Confirming Ammonia Converter Efficiency  
Step No.  
Action  
Comment  
1.  
Press CAL  
The M201E enters the calibrate mode from Sample  
mode.  
2.  
3.  
4.  
Channel Selection Press TNX then Press ENTR.  
Range Selection  
Press LOW range then Press ENTR.  
Press CONC  
The following will be displayed:  
TNx CONV  
Exit  
Press CONV enter 1.0000, Press ENTR  
Press EXIT, EXIT  
05206H DCN5910  
21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Calibration Procedure  
Teledyne API Model M201E NH3 Analyzer Operator Manual  
Table 2-6. NH3 Calibration Procedure – NH3 Gas Through The SAMPLE Port  
Step No.  
Action  
Comment  
1.  
Generate a known Monitor the TNx concentration until it has stabilized;  
level of NH3 gas this signifies that both the NOx and NH3 gases have  
using a calibration stabilized. This could range from 20 minutes to  
gas source  
several hours depending on whether the analyzer /  
calibration system has recently been subjected to  
ammonia gas.  
2.  
NH3 Conversion  
efficiency  
When the TNx channel is stable, note the  
concentration of NH3 displayed. The calculation for  
the NH3 Conversion efficiency is:  
calculation  
(Displayed NH3 concentration) divided by  
(Delivered NH3 concentration)  
Example: The operator is delivering 450 PPB of NH3.  
The displayed concentration for NH3 is 440 PPB then  
the efficiency factor = 440/450 = 0.9777  
3.  
Press CAL  
The M201E enters the calibrate mode from Sample  
mode.  
4.  
5.  
6.  
Channel Selection Press TNX then Press ENTR.  
Range Selection  
Press LOW range then Press ENTR.  
Press CONC  
Since TNx was chosen previously, then the following  
will be displayed:  
TNx CONV  
Exit  
Press CONV enter 0.9777, Press ENTR  
Press EXIT, EXIT  
Note: In the example outlined above, the displayed ammonia  
concentration is being increased by 2.27%. Newer M501 NH3, converters  
should have efficiency values very close to 1.00. Depending on the  
operator’s acceptable level of error, it may not be necessary to change the  
efficiency factor from 1.00. Therefore, calibration with nitric oxide gas is all  
that is required.  
The efficiency factor may also be changed by entering the VARS MENU  
and editing NH3_CE_FACTOR1 = 1.0000.  
22  
05206H DCN5910  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Teledyne API Model M201E NH3 Analyzer Operator Manual  
Calibration Procedure  
Table 2-7. NH3 Calibration Procedure – NH3 Gas through the SPAN Port  
Step No.  
Action  
Comment  
1.  
Press CALS  
The M201E enters the calibrate mode from Sample  
mode. The ammonia gas is supplied through the  
SPAN inlet port on the rear panel of the M501 NH3,.  
Calculate the Efficiency factor as outlined in Steps 2  
in table 2.5.  
2
Table 2.5  
Follow Steps 4 to 6 in table 2.5.  
Note: The response time of the analyzer increases when using the Zero/Span  
valve option.  
05206H DCN5910  
23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Calibration Procedure  
Teledyne API Model M201E NH3 Analyzer Operator Manual  
This page intentionally left blank.  
24  
05206H DCN5910  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3.0 CONFIGURABLE ANALOG OUTPUT OVERVIEW  
There are three different methods to extract concentration data from the E-Series  
analyzers. By one method, the operator can set up an iDas (internal data  
acquisition) configuration either through the user display or using TAPI’s windows  
based software called APICOM. The analyzer stores data internally, which is  
available later for download through the RS-232 or Ethernet port. The second  
method involves the use of hyperlink where the user sends text based commands,  
to retrieve data, through the RS-232 port. With the third method, the operator can  
interface a strip chart recorder and/or data-logger to the analyzer analog outputs.  
The following sections outline the steps necessary to calibrate and configure the  
analog outputs.  
There are four analog outputs on the rear of the analyzer. The default  
configurations of the M201E analog outputs are as follows:  
Table 3-1. Analog Output Data Default Settings  
CHANNEL DEFAULT SETTING  
ANALYZER  
A1  
A2  
A3  
A4  
TNxCNC1  
5 Volts  
NH3CNC1  
5 Volts  
NOCNC1  
5 Volts  
NO2CNC1  
5 Volts  
M201E  
500 PPB  
500 PPB  
500 PPB  
500 PPB  
To access these signals attach a strip chart recorder and/or data-logger to the  
appropriate contacts of the analog output connecter on the rear panel of the  
analyzer.  
ANALOG OUT  
A1  
A2  
A3  
A4  
+
-
+
-
+
-
+
-
Figure 3-1. Analog Output Connector  
05206H DCN5910  
25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Configurable Analog Output Overview  
Teledyne API Model M201E NH3 Analyzer Operator Manual  
Table 3-2. Analog Output Pin-Outs  
PIN  
1
ANALOG OUTPUT  
VOLTAGE SIGNAL  
V Out  
CURRENT SIGNAL  
I Out +  
I Out -  
A1  
2
Ground  
V Out  
3
I Out +  
I Out -  
A2  
A3  
A4  
4
Ground  
V Out  
5
I Out +  
I Out -  
6
Ground  
V Out  
7
N/A  
8
Ground  
N/A  
Additionally A1, A2 and A3 may be equipped with optional 0-20 mA current loop  
drivers. The 4-20 mA option is not available on A4.  
Note: In actuality the analog output configuration of the analyzer may be different  
than stated above. The outputs can be configured differently at the factory  
depending on whether they were assigned during the procurement of the product.  
It is possible to check the configuration of the analyzer by accessing the ANALOG  
I/O CONFIGURATION through the DIAG menu.  
The analyzer operates in Dual Mode during gas detection. This means each of the  
measured concentrations can have two separate slopes and offsets, one for the  
low range and one for the high range. Though uncommon, a user may decide to  
calibrate the analyzer with nitric oxide at, for example, 100 PPB using the LOW  
range and then perform another calibration at 450 PPB using the HIGH range.  
Through the analog outputs the user can then assign analog output A1 to  
TNxCNC1 and output A2 to TNxCNC2. Gases with the “1” designation use the  
slope and offset for the LOW range, while gases with the “2” designation will use  
the slope and offset for the HIGH range. It is recommended that both the LOW and  
HIGH ranges be calibrated at the same time. Independent of whether the HIGH  
range is actually being outputted to the analog outputs.  
26  
05206H DCN5910  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Teledyne API Model M201E NH3 Analyzer Operator Manual  
Configurable Analog Output Overview  
3.1 ANALOG OUTPUT - ELECTRONIC CONFIGURATION  
The final step in configuring the analyzer’s four analog output channels is to set the  
electronic signal type and range of each channel. This consists of:  
Selecting an output type for A1 – A4 (voltage or current if an optional  
current output driver has been installed) and the signal level that matches  
the input requirements of the recording device attached to the channel.  
Calibrating the electronic circuitry of the output channel. For each channel  
this can be done either automatically or manually.  
Adding a bipolar recorder offset to the signal if required.  
In its standard configuration the analyzer’s Analog Outputs can be set for the  
following DC voltage signal levels. Each range is usable from -5% to + 5% of the  
rated range.  
Table 3-3. Analog Output Voltage Range Min/Max  
MINIMUM OUTPUT  
MAXIMUM OUTPUT  
RANGE SPAN  
0-0.1 VDC  
0-1 VDC  
-0.005 VDC  
-0.05 VDC  
-0.25 VDC  
-0.5 VDC  
0.105 VDC  
1.05 VDC  
5.25 VDC  
10.5 VDC  
0-5 VDC  
0-10 VDC  
The default offset for all ranges is 0 VDC.  
The following DC Current output options are also available from the factory:  
Table 3-4. Analog Output Current Loop Min/Max  
RANGE SPAN  
2-20 mADC  
MINIMUM OUTPUT  
1 mADC  
MAXIMUM OUTPUT  
21 mADC  
4-20 mADC  
3 mADC  
21 mADC  
The default offset for all ranges is 0 mADC.  
05206H DCN5910  
27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Configurable Analog Output Overview  
Teledyne API Model M201E NH3 Analyzer Operator Manual  
3.2 ANALOG OUTPUT SIGNAL TYPE AND RANGE SELECTION  
To select an analog output signal type (DC voltage or current) for one of the  
analyzer analog output channels press SETUP - MORE -DIAG - 818…ENTER-  
NEXT…NEXT:  
FROM ANALOG I/O CONFIGURATION MENU  
DIAG  
PREV  
ANALOG I / O CONFIGURATION  
NEXT  
ENTR  
EXIT  
EXIT  
DIAG AIO  
AOUTS CALIBRATED: NO  
Press SET> to select the Analog  
Output channel to be configured:  
< SET SET> CAL  
DISPLAYED AS  
DATA_OUT_1  
DATA_OUT_2  
DATA_OUT_3  
DATA_OUT_4  
=
=
=
=
=
CHANNEL  
A1  
A2  
A3  
A4  
DIAG AIO  
DATA_OUT_2: 5V, NH3CNC1, CAL  
Then Press EDIT to continue  
< SET SET> EDIT  
EXIT  
DIAG AIO  
DATA_OUT_2 RANGE: 5V  
Press SET> to select the channel  
SET> EDIT  
EXIT  
parameter:  
DATA_OUT_2 RANGE: 5V  
DATA_OUT_2 OVERRANGE: ON  
DATA_OUT_2 REC OFS: 0 mv  
DATA_OUT_2 AUTO CAL: ON  
DATA_OUT_2 CALIBRATED: YES  
DATA_OUT_2 OUTPUT: ON  
DATA_OUT_2 DATA: NO2CNC1  
DATA_OUT_2 SCALE: 500 PPB  
DATA_OUT_2 PRECISION: 1  
DATA_OUT_2 UPDATE: 5 SEC  
DIAG AIO  
DATA_OUT_2 RANGE: 5V  
0.1V 1V 5V 10V CURR  
ENTR EXIT  
Press ing ENTR records the new setting  
and returns to the previous menu  
Pressing EXIT ignores the new setting  
Then Press EDIT to continue  
and returns to the previous menu  
These Keys set  
the Signal Level  
of the selected  
channel  
Figure 3-2. Analog Output Signal Type  
28  
05206H DCN5910  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Teledyne API Model M201E NH3 Analyzer Operator Manual  
Configurable Analog Output Overview  
3.3 SELECTING AUTO OR MANUAL ANALOG OUTPUT CALIBRATION  
The analog outputs configured for voltage mode can be calibrated either  
automatically or manually. In its default mode the instrument is configured for  
automatic calibration. Note: manual calibration should only be used for the 0.1V  
range or in cases where the outputs must be closely matched to the  
characteristics of the recording device. Outputs configured for automatic  
calibration can be calibrated as a group or individually.  
To select auto-calibration for a particular channel press the following:  
05206H DCN5910  
29  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Configurable Analog Output Overview  
Teledyne API Model M201E NH3 Analyzer Operator Manual  
FROM ANALOG I/O CONFIGURATION MENU  
DIAG  
ANALOG I / O CONFIGURATION  
Exit to Return  
to the main  
Sample Display  
PREV  
NEXT  
ENTR  
EXIT  
EXIT  
DIAG AIO  
AOUTS CALIBRATED: NO  
Press SET> to select the Analog Output  
channel to be configured:  
< SET SET> CAL  
DISPLAYED AS  
DATA_OUT_1  
DATA_OUT_2  
DATA_OUT_3  
DATA_OUT_4  
=
=
=
=
=
CHANNEL  
A1  
A2  
A3  
A4  
DIAG AIO  
DATA_OUT_2: 5V, NH3CNC1, CAL  
Then Press EDIT to continue  
< SET SET> EDIT  
EXIT  
DIAG AIO  
DATA_OUT_2 RANGE: 5V  
SET> EDIT  
EXIT  
DIAG AIO  
DATA_OUT_2 OVERRANGE: ON  
< SET SET> EDIT  
EXIT  
EXIT  
DIAG AIO  
DATA_OUT_2 REC OFS: 0 mv  
< SET SET> EDIT  
DIAG AIO  
DATA_OUT_2 AUTO CAL: ON  
< SET SET> EDIT  
EXIT  
Toggles the Auto  
Cal Mode  
ON/ OFF for this  
Analog Output  
channel only.  
DIAG AIO  
DATA_OUT_2 AUTO CAL: ON  
ENTR EXIT  
ON  
Pressing ENTR records the new setting  
and returns to the previous menu  
Pressing EXIT ignores the new setting and  
returns to the previous menu  
Figure 3-3. Auto Or Manual Analog Output Calibration  
30  
05206H DCN5910  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Teledyne API Model M201E NH3 Analyzer Operator Manual  
Configurable Analog Output Overview  
To calibrate the output channels as a group press the following, (Note: this is the  
recommended calibration approach):  
FROM ANALOG I/O CONFIGURATION MENU  
Exit at Any  
Time to Return  
to the main  
DIAG  
ANALOG I / O CONFIGURATION  
PREV  
NE XT  
ENTR  
EXIT  
EXIT  
DIAG Menu  
DIAG AIO  
AOUTS CALIBRATED: NO  
< SET SET> CAL  
If AutoCal has been manually  
turned off for any channel the  
message for that c hannel will be  
similar to:  
DIAG AIO AUTO CALIBRATING DATA_OUT_1  
AUTO CALIBRATING DATA_OUT_2  
NOT AUTO CAL DATA_OUT_1  
AUTO CALIBRATING DATA_OUT_4  
Exit to Return  
to the I/O  
DIAG AIO  
AOUTS CALIBRATED:  
YES  
Configuration  
Menu  
< SET SET> CAL  
EXIT  
Figure 3-4. Analog Output Group Calibration  
05206H DCN5910  
31  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Configurable Analog Output Overview  
Teledyne API Model M201E NH3 Analyzer Operator Manual  
3.4 MANUALLY CALIBRATING ANALOG OUTPUT SIGNAL LEVELS  
The analog outputs in voltage mode can be manually calibrated to closely match  
the characteristics of the data recorder. Outputs configured for 0.1V full scale  
should always be calibrated manually.  
Calibration is done through the instrument software in conjunction with a voltmeter  
connected across the output terminals. Adjustments are made using the front  
panel keys. First the zero-point is set then the span-point.  
The software allows this adjustment to be made in 100, 10 or 1 count increments.  
Table 3-5. ZERO/SPAN Analog Output Signal Calibration Tolerances  
Full Scale  
0.1 VDC  
1 VDC  
Adjust Zero Within  
±0.0005V  
Span Voltage  
90 mV  
Adjust Span Within  
±0.001V  
±0.001V  
900 mV  
±0.001V  
5 VDC  
±0.002V  
4500 mV  
4500 mV  
±0.003V  
10 VDC  
±0.004V  
±0.006V  
See Table 6-8 for pin  
assignments on the  
for ANALOG  
VDC  
connector located on  
the instruments rear  
panel  
+DC Grnd  
V OUT +  
V OUT -  
V IN +  
V IN -  
Recording  
Device  
ANALYZER  
Figure 3-5. Setup For Calibrating Analog Output Signal Levels  
32  
05206H DCN5910  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Teledyne API Model M201E NH3 Analyzer Operator Manual  
Configurable Analog Output Overview  
To make these adjustments the auto calibration feature “DATA_OUT_1 AUTO  
CAL: OFF” must be turned off for each output channel then press the following:  
FROM ANALOG I/O CONFIGURATION MENU  
DIAG  
ANALOG I / O CONFIGURATION  
PREV  
NEXT  
ENTR  
EXIT  
EXIT  
DIAG AIO  
AOUTS CALIBRATED: NO  
Press SET> to select the Analog Output  
channel to be configured:  
< SET SET> CAL  
DISPLAYED AS  
DATA_OUT_1  
DATA_OUT_2  
DATA_OUT_3  
DATA_OUT_4  
=
=
=
=
=
CHANNEL  
A1  
A2  
A3  
A4  
DIAG AIO  
DATA_OUT_1 :5V, TNXCNC1, NO CAL  
Then Press EDIT to continue  
< SET SET> EDIT  
EXIT  
DIAG AIO  
DATA_OUT_1 RANGE: 5V  
SET> EDIT  
EXIT  
Press SET Until  
DIAG AIO  
< SET  
DATA_OUT_1 CALIBRATED: NO  
CAL  
EXIT  
These keys  
increment/decrement the  
ZERO/SPAN D-to-A converter  
output by  
DIAG AIO DATA_OUT_1 VOLT–Z : 0 mV  
U100 UP10 UP DOWN DN10 D100 ENTR EXIT  
100, 10 or 1 counts  
respectively.  
Continue adjustments until the  
voltage measured at the  
output of the analyzer and/or  
the input of the recording  
device matches the value in  
the upper right hand corner of  
the display to the tolerance  
listed previously.  
EXIT ignores the  
new setting.  
DIAG AIO DATA_OUT_1 VOLT–S : 4500 mV  
U100 UP10 UP DOWN DN10 D100 ENTR EXIT  
ENTR accepts the  
new setting.  
The analyzer display WILL  
NOT CHANGE. Only the  
voltage reading of your volt  
meter will change.  
DIAG AIO  
< SET  
DATA_OUT_1 CALIBRATED: YES  
CAL EXIT  
Figure 3-6. Analog Output Manual Calibration  
05206H DCN5910  
33  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Configurable Analog Output Overview  
Teledyne API Model M201E NH3 Analyzer Operator Manual  
3.5 ANALOG OUTPUT RECORDER OFFSET ADJUSTMENT  
An offset can be added to individual output channels by pressing the following:  
FROM ANALOG I/O CONFIGURATION MENU  
DI AG  
ANALOG I / O CONFIGURATION  
PREV  
NE XT  
ENTR  
EXIT  
EXIT  
DIAG AIO  
DIAG AIO  
AOUTS CALIBRATED: NO  
Press SET> to select the Analog Output  
channel to be configured:  
SET> CAL  
DISPLAYED AS  
DATA_OUT_1  
DATA_OUT_2  
DATA_OUT_3  
DATA_OUT_4  
=
=
=
=
=
CHANNEL  
A1  
A2  
A3  
A4  
DATA_OUT_2: 5V, NH3CNC1, CAL  
Then Press EDIT to continue  
< SET SET> EDIT  
EXIT  
DIAG AIO  
DATA_OUT_2 RANGE: 5V  
SET> EDIT  
EXIT  
Press SET Until  
DIAG AIO  
DATA_OUT_2 REC OFS: 0 mV  
< S ET SET> EDIT  
EXIT  
DIAG AIO  
RECORD OFFSET: 0 MV  
These Keys set  
the Recorder  
Offset  
+
0
0
0
0
ENTR EXIT  
Pressing ENTR records the new setting  
and returns to the previous menu  
of the s elected  
channel  
Pressing EXIT ignores the new setting and  
returns to the previous menu  
Figure 3-7. Analog Output Recorder Offset Calibration  
34  
05206H DCN5910  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
4.0 MAINTENANCE SCHEDULE  
The maintenance requirements of the M201E are the same as a standard M200E  
NOx analyzer but with additional maintenance items (Table 4-1). Please refer to  
Section 13 in the M200E Operator Manual for the M200E maintenance schedule  
and for maintenance and repair procedures.  
The M201E requires the following additional maintenance items.  
Table 4-1. Preventative Maintenance Schedule  
Item  
Maintenance Interval  
Reference Section  
Section 8.2.3 M200E Manual  
NO2 molybdenum converter  
Test Every Three Months  
M501 NH3 Converter  
Reaction Cell  
AutoZero flow check  
Clean annually or as necessary Section 9.3.9 M200E Manual  
Quarterly as needed Section 4.4 of this manual.  
4.1 M501 NH3 MAINTENANCE  
The M501 NH3 external converter is operated at 825 ºCelsius. Because of this the  
stainless steel ¼” tube internal to the converter slowly oxidizes and may require  
replacement. Over time the converter’s efficiency will degrade, due to aging of  
the internal catalyst. This is characterized by a CONV value on the TNx channel  
of < 0.8 or > 1.2, despite the converter being at temperature. The efficiency of the  
converter should be checked quarterly. The converter should be cleaned out on a  
yearly basis by removing the outer tube and shaking the fine dust particles from  
both the outer and inner tubes. This is also a good time to replace the catalyst  
screen, if necessary.  
The following procedure describes how to disassemble and replace converter  
parts. The various parts in the converter become delicate and brittle after  
prolonged exposure to high temperatures. It is therefore a good idea to have a  
complete set of replacement parts on hand before starting, as listed in Table 4-2.  
05206H DCN5910  
35  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Maintenance Schedule  
Teledyne API Model M201E NH3 Analyzer Operator Manual  
Table 4-2. M501 NH3 Converter Rebuild Parts List  
Part No.  
KIT000193  
KIT000139  
Description  
RETROFIT, M501NH INNER TUBE w/SCREEN  
REBUILD, CERAMIC BUSHINGS, M501NH  
HE0000007 CERAMIC HEATER, 220W@60V M501XX  
4.2 REPLACING THE CATALYTIC CARTRIDGE  
1. Turn off power to the converter and allow to cool. IT IS VERY IMPORTANT  
THAT THE CONVERTER BE COOL BEFORE ATTEMPTING ANY  
DISASSEMBLY OR REPAIRS.  
CAUTION VERY HOT  
Will cause severe burns – disassembly while hot will  
damage other converter components. Allow sufficient  
time to cool.  
2. Remove the chassis cover of the M501 NH3, remove the aluminum cover over  
the oven, remove the two U-shaped hold down clamps.  
3. Disconnect the 1) stainless steel and 2) PTFE tubes from the converter  
cartridge.  
4. Loosen the nut holding the cartridge assembly in the U-shaped angle bracket  
mounted on the chassis floor.  
5. Gently lift and tilt the converter assembly out of the U-shaped bracket and  
slide the assembly out of the oven.  
6. Loosen the ¼” SS tube fitting nearest the hold-down bracket; this fitting holds  
the central tube in the converter. Replace the tube that is part of KIT000193.  
7. Re-assemble the converter by doing the above steps in reverse order.  
Figure 4-1. Catalytic Cartridge  
36  
05206H DCN5910  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Teledyne API Model M201E NH3 Analyzer Operator Manual  
Maintenance Schedule  
4.3 REPLACING THE THERMOCOUPLE  
The following instructions provide the necessary information to replace the  
existing thermocouple. Figure 4-2 is provided for reference.  
You must obtain the following replacement parts kit from T-API:  
010820000 , Type K thermocouple  
You will need the following tools:  
Nutdriver, 5/16  
Nutdriver, 11/32  
Diagonal Cutter  
Philips head Screwdriver #2  
CAUTION  
Avoid damage to the unit: use only Type K thermocouple in  
the M501 NH3, converter. It can be distinguished by its wire  
colors: red and yellow. If you have any other thermocouple,  
do NOT install it; call T-API to obtain the correct part.  
Figure 4-2. Thermocouple Location  
Once you have the right tools and parts, replace the thermocouple as follows:  
1. Ensure power is removed from the M501 NH3, Converter. If the Converter has  
been operational you will need to wait for at least 2 hours for the Converter oven to  
cool before continuing with the replacement of the thermocouple.  
05206H DCN5910  
37  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Maintenance Schedule  
Teledyne API Model M201E NH3 Analyzer Operator Manual  
2. Remove the cover from the Converter chassis.  
3. Unscrew the (4) nuts that secure the front panel to the chassis. They are located just  
behind the Front Panel along the top.  
4. Lower the Front Panel to gain easier access to the end of the quartz tube.  
5. Unscrew the (3) nuts that secure the inner cover protecting the Heater Block and  
quartz tube. Remove this cover.  
6. Cut the tie-wrap that secures the thermocouple to the fitting at the end of the quartz  
tube.  
7. Loosen the Teflon fitting at the end of the quartz tube taking care not to put any  
stress on the tube, and slide the fitting off the tube.  
8. Remove the thermocouple.  
9. Disconnect the thermocouple wires from the Temperature Controller noting that the  
‘Yellow’ wire goes to pin 1 and the ‘Red’ wire goes to pin 2.  
10. The thermocouple should slide into the Heater Block.  
11. Replace the Tie-Wrap Hold-Down.  
12. Form the thermocouple wire so that it rests in the cavity with little movement.  
13. Connect the (2) wires of the thermocouple to the Temperature Controller. The  
“yellow” wire should be connected to pin 1 and the “red” wire should be connected to  
pin 2.  
14. At this point, all connections have been made, both electrically and pneumatically. A  
leak check should be performed on the Converter to verify that all connections are  
leak free. If a leak is detected, the leak should be resolved before continuing.  
15. Install the inner cover of the Heater Block and secure with the (3) nuts. Close the  
Front Panel and secure with the (4) nuts. Install the top cover on the Converter  
chassis.  
16. The Converter is now ready for the application of power. You will be looking for an  
indication from the Temperature Controller that it is functioning correctly and driving  
the heater to the desired “set” temperature. Apply power now.  
17. After the Converter comes to the regulated temperature, perform the Auto Tune  
function to tune the Temperature Controller to the new thermocouple.  
18. After the Auto-Tune process is completed, verify that the “process” temperature is  
indicating that the desired temperature is stable and being regulated.  
The converter is now ready for operation.  
38  
05206H DCN5910  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model M201E NH3 Analyzer Operator Manual  
Maintenance Schedule  
4.4 AUTOZERO FLOW CHECK  
Since the M201E is an ammonia analyzer, the flow through the AutoZero orifice is  
especially important. Check the AutoZero flow as follows:  
1. This procedure should be performed with the sample pump running.  
2. Remove the top cover of the analyzer. Locate the vacuum manifold at the  
center rear of the chassis. Locate the 1/8” tube fitting located on the very left  
side, as viewed from the rear of the analyzer. It will have a label of “0.010”  
indicating the flow orifice installed.  
3. Remove the fitting and attach a calibrated flowmeter to the block fitting capable  
of measuring in the range of 500 cc/min. The flowmeter should indicate a flow  
of 500 cc/min 10%.  
4. If the flow is outside this range (most likely lower, due to plugging):  
1. Turn off the sample pump and instrument.  
2. Remove the 1/8” pipe-to-tube fitting and remove the ¼” sintered filter (p/n  
FL0000001). This filter is meant to protect the orifice; it will usually  
become plugged and need replacement rather than the orifice. Replace  
the filter, and then re-assemble the manifold. Restart the sample pump  
and recheck the flow.  
Table 4-3. Analyzer Flow Check  
M501 NH3, Converter sample flow check  
Note: Connect both the Teflontubes between analyzer and converter  
GAS PORT  
Expected Flow (cc/min)  
“Sample In”  
1000 +/- 10 %  
Analyzer sample flow check  
Note: Disconnect both the Teflontubes between analyzer and converter  
GAS PORT  
“To Conv”  
0
Expected Flow (cc/min)  
1000 or 500, +/- 10 %  
When the analyzer is sampling  
TNx only, the bypass flow is  
measured. Therefore, the flow  
alternates between 500 and  
1000 cc/min. This is the NOx  
channel.  
“From Conv”  
500 +/- 10%  
This is the TNx channel.  
05206H DCN5910  
39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Maintenance Schedule  
Teledyne API Model M201E NH3 Analyzer Operator Manual  
This page intentionally left blank.  
40  
05206H DCN5910  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5.0 ALARMS AND CAUTIONS  
A functional block diagram of the analyzer is shown in Figure 5-1.  
5.1 ALARM  
During initial warm up the internal Zero/Span factory option may exhibit a “Block  
Temperature” warning. This can be cleared once the block temperature stabilizes  
at 50 degrees Celsius. The block temperature is the actual temperature of the IZS  
block manifold inside the M501 NH3, converter. If the block temperature warning  
exhibits a negative number such as (-37), then the umbilical cord is probably not  
attached or the Thermistor has become disconnected.  
5.2 CAUTION  
Be aware that the ammonia converter operates at 825 degrees Celsius. Do not  
flow highly flammable gases through the M201E analyzer. When the analyzer is  
not in use, it is advisable to turn off the M501 NH3,. If the converter hasn’t been  
powered for an extended period of time it is recommend that the analyzer be  
operated overnight prior to use.  
Before spanning the analyzer it is good practice to observe the NORM PMT  
signal. Its value should be equal to twice the concentration of the delivered span  
gas. A common error made during calibration is to span the analyzer with either  
too low a gas concentration or an incorrect value entered into the CONC menu  
for TNx and NOx. Commonly this will cause a slope value much greater than  
one.  
Note: Always zero then span as a basic rule of thumb.  
05206H DCN5910  
41  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Alarms and Cautions  
Teledyne API Model M201E NH3 Analyzer Operator Manual  
KEYBOARD/DISPLAY  
STD  
BUS  
DIGITAL  
ANALOG I/O  
I/O  
CPU  
MOTHERBOARD  
PREAMP  
RS-232  
ANALOG  
OUTPUTS  
OZONE  
DESTRUCT  
THERMOCOUPLE  
RCELL  
PRESSURE  
VACCUM BLOCK  
VACUUM  
PUMP  
PHOTONS  
REACTION  
CELL  
NO/NOx  
VALVE  
OZONE  
CLEANSER  
AUTOZERO  
VALVE  
MOLYBDENUM  
CONVERTER  
OZONE  
GENERATOR  
SAMPLE  
PRESSURE  
NAFION DRIER  
DFU FILTER  
TNx / NOx  
VALVE  
M501NH  
NH3 CONVERTER  
SAMPLE  
MANIFOLD  
Figure 5-1. M201E Functional Block Diagram  
42  
05206H DCN5910  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
6.0 M201E SPARE PARTS LIST  
While this section lists the spare parts for the M201E, it is recommended that you refer to  
our Website or Sales for updates.  
M201E Spare Parts List, Source: PN 06185H (DCN5494)  
Part Number  
000940100  
000940400  
000940500  
000940600  
002730000  
005960000  
005970000  
009690200  
010820000  
011630000  
013140000  
014080100  
016680600  
018080000  
025070000  
025250000  
037860000  
039700100  
040010000  
040030800  
040400000  
041710000  
041800500  
041920000  
042580000  
042680100  
043220000  
043220100  
043940000  
044100000  
045230100  
Description  
CD, ORIFICE, .003 GREEN  
CD, ORIFICE, .004 BLUE  
CD, ORIFICE, .007 ORANGE  
CD, ORIFICE, .010 BROWN  
CD, FILTER, 665NM (KB)  
AKIT, EXP, 6LBS ACT CHARCOAL (2 BT=1)  
AKIT, EXP, 6LB PURAFIL (2BT=1)  
AKIT, TFE FLTR (FL19) ELEM, 47MM, (100)  
ASSY, THERMOCOUPLE, HICON, M501  
HVPS INSULATOR GASKET (KB)  
ASSY, COOLER FAN (NOX/SOX)  
ASSY, HVPS, SOX/NOX  
PCA, O3 GEN DRIVER, NOX, E SERIES  
AKIT, DESSICANT BAGGIES, (12)  
ASSY, IZS VALVE HEATER, M501NH  
ASSY, VALVE/THERM, M501NH  
ORING, TEFLON, RETAINING RING, 47MM (KB)  
HEATER, BAND, TYPE K, DUAL VOLTAGE(KB)  
ASSY, FAN REAR PANEL, E SERIES  
PCA, PRESS SENSORS (2X), FLOW, E (NOX)  
ASSY, HEATERS/THERMAL SWITCH, RX CELL  
ASSY, CPU, CONFIGURATION, "E" SERIES * (USE WITH DOC 053150100)  
PCA, PMT PREAMP, VR, M200E/EM/EH  
ASSY, THERMISTOR, M200E (KB)  
PCA, KEYBOARD, E-SERIES, W/V-DETECT  
ASSY, VALVE (SS), M200E  
TUBING, CERAMIC, .25OD X .156 ID  
THERMOCOUPLE INSULATING SLEEVE, M501NH *  
PCA, INTERFACE, ETHERNET, E-SERIES  
MANUAL, OPERATORS, M200E  
PCA, RELAY CARD, E SERIES  
04550010A ASSY, ORIFICE HOLDER, 4 MIL  
04550030A ASSY, ORIFICE HOLDER, 10 MIL, (NOX) (KB)  
05206H DCN5910  
43  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
M201E Spare Parts List  
Teledyne API Model M201E NH3 Analyzer Operator Manual  
M201E Spare Parts List, Source: PN 06185H (DCN5494)  
Part Number  
Description  
046030000  
AKIT, CH-43, 3 REFILLS  
PCA, SERIAL INTERFACE, w/ MD, E SERIES (USE WITH ACROSSER CPU  
041710000)  
048620200  
049310100  
051211000  
052060000  
052930200  
053150100  
053950000  
055710000  
PCA, TEC DRIVER, PMT, M100/200, E SERIES  
ASSY, O3 DESTRUCT W/FTGS (MNFLD - EXH) *  
MANUAL, 201E STDL ADDENDUM  
ASSY, BAND HEATER TYPE K, M200EX  
DOC, w/SOFTWARE, M201E* (USE WITH ACROSSER CPU 041710000)  
MANUAL, OPERATORS, M200E w/M201E ADDEND  
ASSY, FLT, INLINE SS, 3/8” CONN, 0.5 MIC  
PCA, E-SERIES MOTHERBD, GEN 5-ICOP (ACCEPTS ACROSSER OR ICOP  
CPU)  
ASSY, MOLY GUTS w/WOOL, M101E/M200EX  
PCA, SER INTRFACE, ICOP CPU, E- (OPTION) (USE WITH ICOP CPU  
062870000)  
058021100  
062390000  
062420200  
062870000  
063570100  
CPU, PC-104, VSX-6150E, ICOP *(KB) (MUST USE WITH 058021100 MB)  
DOM, w/SOFTWARE, M201E * (USE WITH ICOP CPU 062870000)  
CH0000037 PURACARB AMMONIA  
CP0000014 CONTROLLER, TEMP, W/PG-08 (CN262)  
DS0000025 DISPLAY, E SERIES (KB)  
FA0000006 FAN, 115VAC (KB)  
FL0000001 FILTER, SS  
FL0000003 FILTER, DFU (KB)  
FL0000034 FILTER, DISPOSABLE, PENTEK (IC-101L)(KB)  
FT0000010 CONNECTOR-ORING, SS, 1/8" (HK)  
HE0000007 CERAMIC HEATER, 220W@60V M501XX  
HW0000005 FOOT  
HW0000020 SPRING  
HW0000030 ISOLATOR, M100/M200 A/E/H/U  
HW0000031 FERRULE, SHOCKMOUNT  
HW0000036 TFE TAPE, 1/4" (48 FT/ROLL)  
HW0000101 ISOLATOR  
HW0000416 COVER, CRYDOM RELAYS, RL9, 19 and 20  
HW0000453 SUPPORT, CIRCUIT BD, 3/16" ICOP  
KIT000095  
KIT000219  
KIT000253  
KIT000254  
AKIT, REPLACEMENT COOLER, A/E SERIES  
KIT, 4-20MA CURRENT OUTPUT (E SERIES)  
ASSY and TEST, SPARE PS37, E SERIES  
ASSY and TEST, SPARE PS38, E SERIES  
OR0000001 ORING, 2-006VT *  
OR0000002 ORING, 2-023V  
OR0000025 ORING, 2-133V  
OR0000027 ORING, 2-042V  
OR0000034 ORING, 2-011V FT10  
44  
05206H DCN5910  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model M201E NH3 Analyzer Operator Manual  
M201E Spare Parts List  
M201E Spare Parts List, Source: PN 06185H (DCN5494)  
Part Number  
Description  
OR0000039 ORING, 2-012V  
OR0000044 ORING, 2-125V  
OR0000083 ORING, 105M, 1MM W X 5 MM ID, VITON  
OR0000086 ORING, 2-006, CV-75 COMPOUND(KB)  
OR0000094 ORING, 2-228V, 50 DURO VITON(KB)  
RL0000015 RELAY, DPDT, (KB)  
RL0000019 SSRT RELAY, TA2410, CE MARK  
RL0000020 SSRT RELAY, TD2410, CE MARK  
SW0000040 PWR SWITCH/CIR BRK, VDE CE (KB)  
SW0000051 SWITCH, POWER CIRC BREAK VDE/CE, w/RG(KB  
SW0000058 SWITCH, THERMAL/450 DEG F  
WR0000008 POWER CORD, 10A  
05206H DCN5910  
45  
Download from Www.Somanuals.com. All Manuals Search And Download.  
M201E Spare Parts List  
Teledyne API Model M201E NH3 Analyzer Operator Manual  
This page intentionally left blank.  
46  
05206H DCN5910  
Download from Www.Somanuals.com. All Manuals Search And Download.  
7.0 CALCULATING MOLY CONVERTER EFFICIENCIES  
7.1 PURPOSE  
To provide instructions on how to calculate the efficiency of a Moly converter  
when using a GPT method of testing converters, by using the US EPA method,  
where the actual concentration of ozone is not a factor in the accuracy of the  
calculation of the converter efficiency. This procedure is based on the Code of  
Federal Regulations, Title 40, Chapter I, subchapter C, Part 50, Appendix F.  
7.2 TOOLS  
API M700 calibrator with O3 Gen option (or equivalent)  
7.3 PARTS  
NONE  
7.4 PROCEDURE  
1.  
For the sake of numbers to input into this note, we have chosen 450 PPB NO  
gas as our reference point, you don’t have to pick these values, they are just an  
example. There is also an assumption that the analyzer has a good calibration  
done @ 450 PPB NO span gas. If this is not the case, then once you are done  
with the leak check on the analyzer input your 450 PPB NO span gas and  
calibrate the analyzer.  
NOTE: for the GPT to be performed correctly, there must be a minimum of 10%  
MORE NO than O3 produced. For example, if the Ozone produced is 400 PPB  
then the NO used must be 440 or more. Typically 450 PPB NO is made and 400  
PPB of Ozone is produced.  
2.  
3.  
Leak check machine to ensure that there are no leaks in the analyzer.  
If you have input a CE factor into the instrument firmware (this would be in the  
CAL-CONC-CONV-SET menu) other than 100%, change this back to 100% for  
the duration of this test.  
4.  
The first gas check is to test to see how much the converter is eating NO gas or  
out gassing NO gas. Bypass the converter in the machine, by placing a short  
piece of tubing in place of the converter. Perform a straight dilution with NO gas  
and air as a diluent gas. Input this 450 PPB NO gas into the analyzer, allow the  
machine to stabilize, and write down the NOx value on your data sheet on line 3.  
05206H DCN5910  
47  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Calculating Moly Converter Efficiencies  
Teledyne API Model M201E NH3 Analyzer Operator Manual  
5.  
Remove the converter bypass and install the converter back into the NOx sample  
stream, such that the NO sample goes through the converter again and allow the  
machine to stabilize. Write down your NOx value on your data sheet on line 4  
AND line 6 of the data sheet.  
6.  
7.  
Note the NO value and input that on line 9 of the data sheet.  
Subtract line 3 from line 4 and write that number down on line 5. The spec on the  
data sheet is the value that we use here in house, and your spec might be a bit  
higher. We have found that on NEW Moly converters this spec is a good one  
that predicts a good performing Moly converter, but in an older converter might  
eat a bit more NO, and this would be acceptable. If it is a constant value, or  
changes little over time, this is not a problem the machine will calibrate this out.  
8.  
The next step is to perform your GPT. Generate the same 450 PPB NO gas and  
input 400 PPB of O3 (or generate 450 PPB NO and 400 PPB NO2, if that’s what  
your calibrator says). Allow the machine to stabilize for 10 minutes and then  
write down the NOx value on line 7 and the NO value on line 10.  
9.  
Subtract line 7 from line 6 and put that onto line 8  
Subtract line 10 from line 9 and put that onto line 11  
10.  
11.  
Put the number from line 8 into the letter A on line 12 and put the number from  
line 11 into the letter B on line 12.  
12.  
13.  
Divide A by B and multiply it by 100 and put it into letter C on line 12.  
Put the number in letter C onto the C on line 13 and subtract that value from 100  
and put it into letter D on line 13. this is the converter efficiency.  
14.  
This value should be >96%. For CEMS applications, a CE of <96% might be  
acceptable, depending on application and the guideline set up by the regulatory  
agency. In any application, check with your regulatory agency to see what the  
minimum CE factor is before replacing the converter.  
48  
05206H DCN5910  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MOLY TEST DATA SHEET  
Line #  
TEST  
RESULT  
2
3
4
LEAK-CHECK (WHEN HOT)  
YES / NO  
NOX RESPONSE (MOLY BYPASSED)  
NOX RESPONSE (MOLY IN-LINE)  
__________  
__________  
5
OUT-GASSING / EATING (NO – NOX)  
__________ (>-5 PPB, <5  
PPB)  
6
7
(NOx ORIG  
)
(NOX mode, O3 off)  
(NOX mode, O3 on)  
__________ PPB  
__________ PPB  
(NOx REM  
)
8
NOX LOSS (9A - 10B)  
__________ (A) (<4% of  
NOx ORIG;  
ex: for  
450PPB  
4% is 18PPB)  
9
(NO ORIG  
)
(NO mode, O3 off)  
(NO mode, O3 on)  
__________ PPB  
__________ PPB  
10  
11  
(NO REM  
)
NO2 (9B - 10A)  
(>300PPB)  
__________ (B)  
12  
Efficiency LOSS [ ( A / B ) x 100 ] = [ ( ____A____ / ____B____ ) x 100 ] =  
____C____%  
13  
Total Conv Eff [ 100% – C ] = 100% - ____C_____ = _____D_____ % ( > 96%)  
05206H DCN5910  
49  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Toshiba CRT Television CN36Z71 User Manual
Trane Computer Keyboard TCONT600AF11MA User Manual
Tricity Bendix Cooktop SG550 1 User Manual
Tripp Lite Surge Protector TLP31SAT User Manual
Tyco Electronics Computer Monitor 3239L User Manual
Uniden Cordless Telephone XC800 User Manual
Viking Range F60067 User Manual
VocoPro Karaoke Machine DA 5800RX User Manual
VTech Handheld Game System A DAY IN THE LIFE OF SPONGE User Manual
Wayne Septic System 330502 001 User Manual