Teledyne Oxygen Equipment AX300 I User Manual

Portable Oxygen Analyzer  
OPERATING & SERVICE INSTRUCTIONS FOR  
AX300-I  
PORTABLE OXYGEN ANALYZER  
P/N M75708  
REV 0  
ECO # 03-0038  
0086  
TYPE B EQUIPMENT:  
Equipment providing a particular degree  
of protection against electric shock,  
particularly regarding—  
• Allowable LEAKAGE CURRENT  
• Reliability of the protective earth  
connection (if present).  
Teledyne Analytical Instruments  
i
Download from Www.Somanuals.com. All Manuals Search And Download.  
Portable Oxygen Analyzer  
About This Manual  
The AX300-I Operator's Manual provides both introductory and  
detailed information for configuring and operating these instruments.  
The manual takes you from the time you unpack the instrument until you  
complete the first gas analysis. The bulk of the manual contains  
operating procedures and information. There are also cautions,  
warnings, and guidelines to ensure that your analyzer operates normally  
and to its full potential. A troubleshooting section is available to assist  
you with common problems and a complete product specifications and  
spare parts list is included as an appendix.  
Chapter 1: An introduction to the analyzer and its  
components, features and applications.  
Chapter 2: Step-by-step set-up procedures and  
information.  
Chapter 3: A guide for daily operational maintenance and  
troubleshooting.  
Appendix:  
Specifications and available spare part options  
for the analyzer, and detailed application  
considerations to aid in troubleshooting, etc.  
How To Use This Manual  
This manual is designed to walk you through the initial set-up of  
the AX300 Portable Oxygen Analyzer. After you have used it to initially  
install your analyzer, it becomes a quick reference guide to help you  
with specific questions or operating problems.  
Before you turning on the instrument, you are advised to read the  
safety information on the next few pages and the information found in  
Chapters 1 and 2. These chapters acquaint the user with the instruments  
use and operation before placing it into operation.  
Teledyne Analytical Instruments  
iii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
AX300-I  
Safety Messages  
Your safety and the safety of others are very important. Please  
carefully read the following safety messages.  
Safety message are indented to alert the user of potential hazards.  
Each safety message is associated with a safety alert symbol. These  
symbols are found in the manual and on the instrument. The definition  
of these symbols is described below:  
CAUTION: Refer to the instructions for details on the  
specific danger. These caution symbols warn of specific  
procedures, which if not followed could cause bodily Injury,  
and/or damage the instrument.  
WARNING: This symbol is use to alert the operator of a  
condition that could cause bodily harm.  
NOTE: Additional information and comments regarding a  
specific component or procedure are highlighted in the form  
of a note.  
CAUTION:  
THE ANALYZER SHOULD ONLY BE USED FOR THE  
PURPOSE AND IN THE MANNER DESCRIBED IN  
THIS MANUAL.  
IF YOU USE THE ANALYZER IN A MANNER OTHER  
THAN THAT FOR WHICH IT WAS INTENDED,  
UNPREDICTABLE BEHAVIOR COULD RESULT  
POSSIBLY ACCOMPANIED WITH HAZARDOUS  
CONSEQUENCES.  
Teledyne Analytical Instruments  
iv  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Portable Oxygen Analyzer  
Table of Contents  
Safety Messages ..........................................................................iv  
List of Figures..............................................................................vii  
List of Tables ..............................................................................viii  
Introduction ...................................................................................9  
1.1 Applicable Standards  
1.2 Features  
10  
10  
11  
12  
12  
12  
13  
1.3 Options  
1.4 Applications  
1.5 Theory of Operation  
1.5.1 Sensor  
1.5.2 Signal Processing  
Operation .....................................................................................14  
2.1 Setup  
14  
15  
16  
17  
17  
17  
19  
21  
22  
22  
24  
24  
24  
24  
25  
2.1.1 Sensor Installation or Replacement  
2.1.2 Mounting  
2.1.2.1 V-Mount Adapter Installation  
2.1.2.2 Universal Mounting Clamp Installation  
2.1.3 Battery Installation  
2.1.4 Calibration  
2.1.5 Output 0-1 VDC or RS232  
2.2 Use  
2.2.1 Procedure  
2.3 Gas Sampling  
2.3.1 Humidity  
2.3.2 Temperature  
2.3.3 Pressure  
2.3.4 Discrepancy in Readings  
Teledyne Analytical Instruments  
v
Download from Www.Somanuals.com. All Manuals Search And Download.  
AX300-I  
25  
2.3.5 Anesthetic Gases  
2.3.5.1 Gases That Induce Reading Error  
2.3.5.2 Care After Use in Nitrous Oxide  
2.3.6 Cleaning  
25  
27  
28  
28  
2.4 Do’s and Don’ts  
Service Manual............................................................................ 31  
3.1 General Service Information  
3.2 Overall Maintenance  
3.3 Battery Maintenance  
3.4 Sensor Maintenance  
3.5 Calibration  
31  
31  
31  
32  
32  
33  
33  
33  
33  
34  
34  
34  
36  
36  
39  
40  
41  
3.6 Gas Sampling  
3.6.1 Humidity  
3.6.2 Temperature  
3.6.3 Pressure  
3.6.4 Discrepancy in Readings  
3.6.5 Anesthetic Gases  
3.6.5.1 Gases That Induce Reading Error  
3.6.5.2 Care After Use in Nitrous Oxide  
3.7 Troubleshooting  
3.8 Watchdog  
3.9 Other Problems with the Instrument  
3.10 Return Authorization for Service  
Appendix...................................................................................... 41  
A.1 Specifications  
41  
43  
43  
A.2 Spare Parts List  
A.3 Optional Accessories  
Index............................................................................................. 45  
Teledyne Analytical Instruments  
vi  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Portable Oxygen Analyzer  
List of Figures  
Figure 1-1: AX300 Front View .........................................................9  
Figure 2-1: Installing the R17MED Sensor ....................................15  
Figure 2-2: Sensor Cable Connection to Analyzer .......................16  
Figure 2-3: Mounting the Sensor in the Tee Adapter.....................16  
Figure 2-4: V-Mount Adapter Installation.......................................17  
Figure 2-5: Brass Insert for Universal Mounting Clamp.................17  
Figure 2-6: Installing Batteries.......................................................18  
Figure 2-8: 0-1 VDC or RS 232 Digital Output Port.......................21  
Teledyne Analytical Instruments  
vii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
AX300-I  
List of Tables  
Table 2-1: Oxygen Reading Error in a Mixture of Anesthetic Gas. 27  
Table 3-1: Oxygen Reading Error in a Mixture of Anesthetic Gas. 35  
Table 3-2 Troubleshooting ............................................................ 36  
Table 3-3 Error Codes................................................................... 39  
Teledyne Analytical Instruments  
viii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Portable Oxygen Analyzer  
Introduction  
Introduction  
The Teledyne Analytical Instruments AX300-I Oxygen Analyzer  
here after referred to as AX300-I is a portable instrument that provides  
fast and accurate oxygen analysis. These instruments are designed to  
analyzer up to 100% oxygen concentration in medical gas mixtures.  
Because they are microprocessor-based, the AX300-I analyzers have a  
unique combination of features that make them very easy to use. The  
operator interface is accomplished through a series of buttons located  
conveniently on the front face of the instrument. The AX300 is shown  
in Figure 1-1.  
Figure 1-1: AX300 Front View  
The LCD display consists of up to 3.5 characters plus a decimal  
point indicator capable of displaying up to 105%. (See Section 1.3  
Options for alternate display configuration). An integral battery life  
indicator is displayed continuously on the AX300-I.  
Teledyne Analytical Instruments  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
Introduction  
AX300-I  
The instrument is powered by three AA alkaline batteries and is  
designed to operate for 2000 hours in Non-alarm State.  
Oxygen analysis is linear across the single range of 0-100% using  
Teledyne’s class R17MED oxygen sensor. A unique sensor failure  
alarm is incorporated which warns the user if the sensor signal is lost or  
low. When this occurs, the SENSOR display flashes. The alarm buzzer  
can be silenced by pressing the ALARM SILENCE key () or  
pressing the unlock key.  
1.1 Applicable Standards  
The AX300-I is built to meet or exceed regulatory and industry  
standards for use as a medical device. These instruments are designed,  
built and tested with the following applicable standards:  
ASTM F 1462: Specifications for Oxygen Analyzers  
ASTM F 1463: Specifications for Alarm Signals  
ISO 7767:  
Oxygen Analyzers for Analyzing patient  
Breathing Mixtures  
ISO 9703-1:  
ISO 9703-2:  
Anesthesia and Respiratory Care Alarm Signals  
Part 1  
Anesthesia and Respiratory Care Alarm Signals  
Part 2  
EN/IEC 60601-1-2: Medical Electrical Equipment—Part 1 General  
Requirements for Safety. Electromagnetic  
Compatibility Requirements and Test  
MIL-STD-810E: Environmental Test Methods  
EN/IEC 60601-1: Medical Electrical Equipment-General  
Requirements for Safety  
1.2 Features  
The AX300-I is a compact, versatile instrument capable of rapidly  
measuring the oxygen content of an atmosphere or environment  
accurately to ±2% over the range 0-100% oxygen. The following  
features are standard on the AX300-I:  
Large easy to read 3 1/2 digit LCD display (see options)  
Automatic LCD back lighting upon key press  
Teledyne Analytical Instruments  
10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Portable Oxygen Analyzer  
Introduction  
Microprocessor controlled  
Sensor fail/disconnect alarm indicator (audible and visual)  
Alarm silence button  
2000 operating hours from 3 AA alkaline batteries  
Battery status indicator  
Stand for upright tabletop deployment  
Hardware for pole clamping and V block support  
Rugged high impact ABS construction  
Splash resistant case.  
Long life (36 months in air) class R17MED sensor  
0-1 VDC digital output (optional RS-232)  
FDA approved and cleared for CSA/CE marking  
1.3 Options  
The following instrument options are available for the AX 300-I  
analyzer:  
A-Option—3-digit LCD display instead of 3 1/2 digit  
B-Option—RS 232 digital output instead of 0-1VDC  
Note: Contact the factory for retrofitting an existing instrument for  
3 digit LCD display. For RS-232 reconfiguration, see  
Section 2.1.5.  
In addition to the above instrument configuration options, the  
following optional equipment is available for your instrument:  
Universal Pole Mounting Clamp (P/N CP 2343)  
V-Mount Pole Clamp (P/N CP 2344)  
V-Mount Wall Adapter P/N B 647)  
0-1 VDC Interface Cable (P/NB-75554)  
RS 232 Interface Cable (P/N B-75555)  
Teledyne Analytical Instruments  
11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Introduction  
AX300-I  
1.4 Applications  
The AX300-I analyzers is intended for spot checking concentration  
of oxygen in a gas mixtures used in medical applications such as  
Anesthesia, Respiratory therapy and is intended for adult, pediatric and  
Neonatal populations.  
These instruments may be used in verifying oxygen concentrations  
in gas mixtures used in:  
Anesthesia  
Respiratory Therapy  
Neonatal Care  
1.5 Theory of Operation  
The AX300-I analyzer can be divided into two major functional  
groups:  
R17MED Oxygen Sensor  
Signal Processing  
The analyzer uses Teledyne Analytical Instruments Patented  
R17MED oxygen sensor. The millovolt output signal from the sensor is  
fed into the electronic signal processor, where it is used to calculate the  
oxygen gas concentration and display it on the LCD screen. There is no  
concentration alarm in the AX300-I analyzer however a unique sensor  
fail/disconnect alarm is incorporated to warn the user of a sensor  
problem. TheSENSOR indicator is illuminated on the LCD and the  
audible and visual alarms are activated whenever a fault is detected.  
1.5.1 Sensor  
The AX300-I uses the Teledyne Class R17MED disposable oxygen  
sensor. The sensor is made up of a sensing cathode and anode (fuel)  
immersed in electrolyte and packaged in a small plastic container.  
Oxygen entering the sensor reacts with the anode and a proportional  
current is collected at the sensing cathode, which is sent to the  
electronics where it is converted into a digital signal and displayed on  
the LCD Screen.  
Attached to the R17MED sensor is a removable plastic diverter.  
This diverter is used to facilitate the transport of gas mixtures through  
Teledyne Analytical Instruments  
12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Portable Oxygen Analyzer  
Introduction  
the sensor. The diverter, packaged separately when shipped, is necessary  
when the tee adapter is used to sample gas flowing through a tube.  
The diverter is not necessary and should not be used when the  
sensor is placed directly in a chamber, or when the sensor is used in  
confined volume analysis, such as incubators and inhalation tents.  
CAUTION:  
THE R17MED SENSOR CONTIANS A CAUSTIC  
ELECTROLYTE AND LEAD. DO NOT TRY TO OPEN  
THE SENSOR ASSEMBLY. CHECK THE SENSOR  
REGULARLY FOR LEAKS. IF THE SENSOR IS  
LEAKING, REPLACE IT. DO NOT TRY TO REPAIR IT.  
CONTACT TELEDYNE FOR THE MATERIAL SAFETY  
DATA SHEET RELATED TO HANDLING AND  
DISPOSAL.  
CAUTION:  
REMOVE AND SAVE THE DIVERTER WHEN THE  
SENSOR IS USED IN CONFINED VOLUME  
APPLICATIONS.  
1.5.2 Signal Processing  
The electrical voltage developed in the sensor is sent to the electronics.  
Processing includes amplification, conversion to digits, and comparison to  
alarm set points if appropriate. Using a microprocessor allows for easier  
setting of alarms, automatic calibration, and self-diagnosis.  
The oxygen level is calculated and then displayed on the liquid  
crystal display (LCD) on the front panel.  
The audio alarm is triggered during a sensor disconnect alarm  
condition. Pressing the ALARM SILENCE key () will provide an  
audible alarm override.  
BLANK  
Teledyne Analytical Instruments  
13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
AX300-I  
Operation  
Note: Upon receipt, INSPECT THE ENTIRE UNIT FOR  
DAMAGE. Check the unit and all included accessories for  
broken or loose parts. If damaged, DO NOT USE. Notify  
the shipper, and consult Teledyne Analytical Instruments.  
Note: This equipment is internally powered using 3 AA batteries.  
CAUTION:  
THE AX300-I, OXYGEN SENSOR AND ASSOCIATED  
HARDWARE ARE NON-STERILE DEVICES. DO NOT  
AUTOCLAVE THE INSTRUMENT OR SENSOR, AS  
THIS WILL DAMAGE THE EQUIPMENT.  
2.1 Setup  
The AX300-I Portable Oxygen Analyzer is suitable for use in many  
medical applications. The unit is equipped with a stand and can be used  
on a tabletop or wall mounted using mounting bracket. An optional pole  
mount clamp is also available.  
To set up and use your AX300-I analyzer:  
1. Install the sensor.  
2. Install the batteries.  
3. Calibrate the unit.  
The control keys are designed for easy operation. A LOCK/UNLOCK  
key has been supplied to prevent accidental changes to critical settings. This  
eliminates unwanted changes in calibration or alarm settings from accidental  
touching or bumping of the keys. To further reduce the possibly of incorrect  
adjustments at least two keys must be pressed in order to modify a critical  
calibration set point value.  
Note: The ALARM SILENCE () key continues to operate  
normally when the lock feature is activated.  
Teledyne Analytical Instruments  
14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Portable Oxygen Analyzer  
Operation  
2.1.1 Sensor Installation or Replacement  
Note: The R17MED oxygen sensor must be installed before the  
oxygen analyzer/analyzer can be operated  
1. Remove the new sensor from its protective bag. Inspect the  
sensor for damage or electrolyte leakage. If the sensor is  
damaged, obtain a replacement. Do not use the defective  
sensor as it may damage the unit.  
WARNING: THE SENSOR ELECTROLYTE IS CAUSTIC. DO NOT  
LET IT COME IN CONTACT WITH SKIN. IF IT DOES,  
FLUSH AFFECTED AREA WITH WATER. DO NOT  
ATTEMPT TO OPEN OR REPAIR THE SENSOR.  
WARNING: THE SENSOR ALSO CONTAINS LEAD. LEAKING OR  
EXHAUSTED SENSORS SHOULD BE HANDLED AND  
DISPOSED OF IN ACCORDANCE WITH LOCAL  
REGULATIONS. CONTACT TELEDYNE FOR THE  
MATERIAL SAFETY DATA SHEET  
2. Plug one end of the coiled cable into the jack receptacle on the  
back end of the R17MED sensor and secure in place with the  
capture nut located at the base of the connector. See Figure 2-1.  
Figure 2-1: Installing the R17MED Sensor  
3. Plug the other end of the coiled cable into the receptacle located on  
the right side of the unit and secure it in place using the capture nut.  
See Figure 2-2.  
Teledyne Analytical Instruments  
15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
AX300-I  
Figure 2-2: Sensor Cable Connection to Analyzer  
Note: When the AX300-I instrument is used for diffusion  
sampling (i.e., incubators, tents, etc.), the plastic flow  
diverter must be removed from the R17MED sensor. If the  
sensor is used in breathing circuits, etc, the diverter must  
be used as shown in Figure 2-3.  
Figure 2-3: Mounting the Sensor in the Tee Adapter  
2.1.2 Mounting  
The AX300-I can be mounted in several ways depending on the  
optional equipment ordered at the time of purchase. See Section 1.3.  
Teledyne Analytical Instruments  
16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Portable Oxygen Analyzer  
Operation  
2.1.2.1 V-MOUNT ADAPTER INSTALLATION  
The V-Mount Adapter consists of a matching plastic plate with  
integral V-grooves that attach to the rear of the instrument.  
To install V-Mount Adapter remove battery compartment door by  
prying up the hinged latch at the bottom of the cover, then slide the  
adapter plate into grooves provided in rear case. Replace the battery  
compartment door and secure door latch. See Figure 2-4.  
Note: The door latch is a tight fit onto the battery cover. Use a  
coin to gently pry up the latch.  
Figure 2-4: V-Mount Adapter Installation  
2.1.2.2 UNIVERSAL MOUNTING CLAMP INSTALLATION  
The Universal Mounting Clamp is supplied with a 1/4-20 screw for  
securing the clamp to the rear of the instrument. A threaded brass insert is  
installed on the back of the instrument for this purpose. See Figure 2-5.  
Figure 2-5: Brass Insert for Universal Mounting Clamp  
2.1.3 Battery Installation  
Note: Three “AA” alkaline batteries must be installed in the unit  
before the analyzer will operate. The unit must be  
recalibrated whenever new batteries are installed.  
To install the batteries:  
Teledyne Analytical Instruments  
17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
AX300-I  
1. Turn the unit off (if it is on).  
2. Hold the instrument face down in the palm of your hand. Use  
a coin to pry up the latch that secures the battery  
compartment door. Remove the battery compartment door.  
CAUTION:  
IMPROPER INSTALLATION OF THE BATTERIES  
MAY RESULT IN DAMAGE TO THE UNIT AND  
BATTERIES.  
Note: Use alkaline batteries only. Other battery types will  
produce erroneous battery test readings.  
3. Install 3 “AA” alkaline batteries into the holder as shown in  
Figure 2-6. Each battery has its own slot. To insure proper  
polarity, place the bottom (flat) or negative end of the battery  
in the end of the holder marked “–”. Place the top (button) or  
positive end of the battery in the end of the holder marked  
“+”. Do this for each battery.  
Figure 2-6: Installing Batteries  
4. Re-install the battery compartment door. When the unit is  
first turned on the display will momentarily display all LCD  
segments. During this period diagnostic tests are being  
conducted to insure the circuits are functioning correctly.  
The unit will activate the audible and visual alarms for about  
1 second. The LCD will flash continuously indicating the  
unit is in the unlocked position ready for calibration.  
Note: When batteries are first installed or power is lost for any  
reason the instrument defaults to the calibration mode. All  
Teledyne Analytical Instruments  
18  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Portable Oxygen Analyzer  
Operation  
keys except the CAL and ON/OFF (I/O) keys are  
inoperable until a successful calibration is achieved.  
5. Note the battery display located below the oxygen readout  
display. It continuously shows a bar graph of the remaining  
life to the batteries. In the case of fresh batteries, it should  
illuminate all 5 segments from the left to the right end of the  
bar. The bar graph indicator is on continuously whenever the  
instrument is powered on.  
2.1.4 Calibration  
The AX300-I should be calibrated before each use and every 8  
hours to maintain accuracy.  
Whenever new batteries are installed or  
removed for any reason, the oxygen analyzer  
defaults to the calibration mode with the LCD  
display flashing 00.0. Only the CAL and  
ON/OFF (I/O) keys are functional at this point.  
Note: For first time and for routine  
maintenance calibrations, make sure  
the sensor and sensor cable are  
installed correctly before attempting  
to calibrate the instrument  
To calibrate the instrument (See Figure 2-7):  
1. Turn the unit on by pressing the ON/OFF (I/O) key.  
Figure 2-7: Calibration S  
2. Check the battery status via the bar graph.  
3. If the LCD is not flashing, press the LOCK/UNLOCK key to  
unlock the keys. LCD will flash indicating changes can be  
made to the settings. If the batteries have just been installed  
the LCD will flash 00.0  
4. As with most oxygen analyzer(s) the highest level of  
accuracy is achieved when calibration is conducted using  
100% oxygen. After installing the flow diverter as noted in  
Section 2.1.1, insert the sensor into the plastic tee and  
connect to a supply of pure dry oxygen flowing at 1-2 liters  
per minute.  
Note: An accessory calibration assembly (P/N C53790) is  
available from Teledyne for use with the R17MED sensor  
Teledyne Analytical Instruments  
19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
AX300-I  
5. Wait about 20 seconds to insure the sample line is completely  
purged with the calibration gas. Press the CAL key. The  
LCD will count down from 9 to 0. During this time the  
microprocessor is measuring the sensor output to determine  
the gas concentration and selects the calibration range i.e.  
100% or 20.9%. When the calibration is complete the LCD  
will display the gas value. Press the LOCK/UNLOCK key to  
save the calibration data.  
Note: The AX300-I can only be calibrated using 100% oxygen or  
room air 20.9%. Improper calibration or use of other gas  
concentrations will activate the SENSOR indicator. To  
repeat the calibration press the LOCK/UNLOCK key and  
press the CAL key.  
6. Remove the sensor from the oxygen supply and confirm the  
LCD reads less than 22% in room air. It is not necessary for  
it to read exactly 20.9%.  
7. It is important to perform the calibration carefully and  
thoroughly, using calibration gases that are free from  
contaminates. Wait for a stable reading before locking in  
calibration point. The accuracy of the instrument is only as  
good as the procedure used to calibrate it.  
Note: A single point air calibration is not recommended unless  
the sensor can be exposed to a known source of fresh  
outdoor air. Hospital room air is often enriched with  
excess oxygen, which will introduce errors into the  
calibration. Air calibration should only be used for  
analyzing oxygen levels between 21% and 40% and  
should never be used where a high degree of accuracy  
is needed.  
Note: Never calibrate the unit in humidified gas, as water vapor  
makes the oxygen concentration appear lower than it really  
is. See Section 2.3.1: Humidity.  
CAUTION:  
DO NOT ADJUST THE CALIBRATION SETTINGS IN  
AIR AFTER THE 100% CALIBRATION, AS THIS WILL  
CANCEL THE MORE ACCURATE 100%  
CALIBRATION. THE 100% CALIBRATION MAY BE  
REPEATED AS MANY TIMES AS DESIRED.  
8. Press the LOCK/UNLOCK key to hold settings. The unit is  
now ready for use.  
Teledyne Analytical Instruments  
20  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Portable Oxygen Analyzer  
Operation  
2.1.5 Output 0-1 VDC or RS232  
The AX300-I Portable Oxygen Analyzer provides signal outputs  
for use with recorders and computers. These instruments are supplied  
standard with a 0-1 VDC output. An optional 0-1 VDC Interface Cable  
(P/N B-75554) is available from Teledyne for this purpose.  
To connect the analyzer to an analog recording device:  
1. Insert one end of the interface cable into the output port on  
the side of the instrument. See Figure 2-8.  
2. Insert the other end into the analog recorder device. Make  
sure the device is equipped to handle a 0-1 VDC signal.  
When properly calibrated, the output signal generated by the  
analyzer is linear and proportional to the oxygen concentration.  
If you requested Option-B (RS 232 digital output) at the time of  
purchase, a digital RS 232 signal is output from the output port shown in  
Figure 2-8. Use the optional RS 232 Interface Cable (P/N B-75555)  
available from Teledyne for connection to a standard RS 232 port on a  
computer or other suitably equipped digital device.  
Figure 2-8: 0-1 VDC or RS 232 Digital Output Port  
If your instrument is set for analog (0-1 VDC) output, you can  
reconfigure it to use the digital output by changing a jumper on the  
internal PC board.  
To activate the digital output:  
Teledyne Analytical Instruments  
21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
AX300-I  
1. Remove the batteries and remove the five screws that hold the  
case together.  
2. Remove the rear case section leaving the PCB in the front half of  
the case.  
3. Remove jumper at position JP3 and reinstall it at position JP7.  
4. Replace the rear cover and secure in place with five screws.  
Install the batteries and recalibrate per Section 2.1.3.  
CAUTION:  
RECORDER/RS232 OUTPUT SIGNAL SHOULD ONLY  
BE CONNECTED TO AN EN60601-1/IEC60606-1  
APPROVED DEVICE.  
To reconfigure the analyzer from a digital (RS 232) output to  
analog (0-1 VDC) output, use the same procedure except in step 3  
remove the jumper from JP7 and replace it at JP3.  
2.2 Use  
2.2.1 Procedure  
Note: Prior to use, always test the batteries. Also check  
calibration, the sensor for leaks and damage, and the  
alarm settings.  
The AX300-I instrument can be used to measure a gas mixture for  
oxygen in two basic modes:  
In the inhalation side of breathing circuit ahead of  
antibacterial filters, humidifiers and medicating devices or  
other instances where gases are flowing to a patient in  
breathing circuits.  
In confined volumes such as incubators or tents.  
When analyzing for oxygen in breathing circuits, the flow diverter  
must be used. The diverter should be screwed onto the threaded front  
end of the R17MED sensor. A tee adapter (plastic, P/N A268, or metal,  
P/N A283) should be placed into the circuit, and the above sensor  
assembly plugged into the tee adapter. See Figure 2-3.  
CAUTION:  
CHECK THE BREATHING CIRCUIT FOR LEAKS. BE  
CERTAIN THAT THE CIRCUIT DOWNSTREAM OF  
THE SENSOR DOES NOT PRODUCE ANY  
Teledyne Analytical Instruments  
22  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Portable Oxygen Analyzer  
Operation  
BACKPRESSURE OR RESTRICTION TO FLOW.  
ERRORS IN READINGS WILL RESULT IF THIS IS  
NOT FOLLOWED.  
THE OXYGEN SENSOR IS A NON-STERILE DEVICE  
AND SHOULD BE USED IN CONJUCTION WITH AN  
ANTIBACTERIAL FILTER. ALWAYS INSTALL THE  
SENSOR ON THE INSPIRED LINE AHEAD OF  
FILTERS, HUMIDIFIERS AND MEDICATING DEVICES.  
NEVER INSTALL THE SENSOR IN A LOCATION  
THAT WILL EXPOSE THE SENSOR TO PATIENTS  
EXHALED BREATH OR SECRETIONS UNLESS YOU  
INTEND TO DISPOSE OF THE SENSOR AND FLOW  
ADAPTER AFTER USE.  
When analyzing for oxygen in confined volumes such as  
incubators, hoods, etc., the flow diverter must be removed from the  
R17MED sensor so that it does not interfere with the rapid exchange of  
gases to and from the sensing surface of the sensor.  
CAUTION:  
FAILURE TO REMOVE THE DIVERTER IN THESE  
APPLICATION AREAS WILL RESULT IN A MARKED  
LOWERING OF THE RESPONSE TIME OF THE  
SENSOR.  
The R17MED sensor can be placed or hung inside incubators, tents,  
etc. When it is necessary to thread the cable through a small hole in  
order to gain access to the inside of a chamber, the cable should be  
disconnected at the sensor, threaded through the hole, and reconnected  
inside the chamber.  
The LOCK/UNLOCK key can be used to lock out any accidental  
interference to the front panel keys. The LOCK/UNLOCK key acts as a  
toggle, pressing LOCK/UNLOCK once renders inactive all keys except  
the ALARM SILENCE () key. Pressing LOCK/UNLOCK a second  
time unlocks the keypad.  
Teledyne Analytical Instruments  
23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
AX300-I  
2.3 Gas Sampling  
2.3.1 Humidity  
Humidity does not directly affect the accuracy of the sensor's  
measurement. However, when a nebulizer or other device is used to  
increase moisture levels in gas mixtures, the moisture actually dilutes  
the mixture. This dilution effect decreases the oxygen concentration.  
For example, if an 80% oxygen gas mixture is humidified to  
saturation at room temperature, the resulting gas mixture will contain  
only 77.5% oxygen. Your portable oxygen analyzer accurately measures  
decreases in the oxygen concentration due to the dilution effects of  
moisture added to gas mixtures.  
As with all oxygen sensors, excessive condensation on the sensing  
surface of the R17MED will block the diffusion of oxygen to the sensor,  
rendering it inoperative. We recommend installing the sensor on the dry  
side of the breathing circuit at all times.  
2.3.2 Temperature  
The R17MED oxygen sensor adjusts for ambient temperature  
changes in the range of 0–40°C (32–106°F). Since the thermistor that  
compensates for these changes is located in the rear of the sensor  
assembly, it is important that gas mixtures, flowing over the front of the  
sensor, be at room temperature. Reading errors may occur if hot gases  
from a heated humidifier are directed past a sensor teed into a breathing  
circuit.  
A small thermal tracking error may be encountered in application  
areas where the entire sensor assembly is placed in the gas mixture to be  
analyzed (e.g., incubators). Holding the sensor in your hand for more  
than a few minutes can also affect the temperature tracking which  
appears as a slow drift on the LCD. No adjustments should be made  
during this period since this error will be eliminated when both the  
thermistor and sensing electrode have had sufficient time to come to  
thermal equilibrium. This can take up to 2 hours.  
2.3.3 Pressure  
Virtually all gas sensors and analyzers measure the partial pressure,  
not the percentage, of the gas that they sense. The only time that these  
instruments can accurately read percentages is when the total pressure  
does not vary over time between calibrations and use. For this reason it  
is important to calibrate the AX300-I oxygen sensor at regular intervals.  
Teledyne Analytical Instruments  
24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Portable Oxygen Analyzer  
Operation  
It is recommended that the unit be calibrated prior to each use or every 8  
hours.  
When the sensor is connected to a ventilator circuit, the alternating  
“breathing” pressure cycles generated by the ventilator will be sensed as  
an increase in the oxygen percentage (especially if the sensor is fast  
enough to sense the changes, as is the R17MED). In reality, the  
percentage of oxygen is not changing; it is the total pressure that is  
increasing producing a corresponding increase in the partial pressure of  
oxygen. A hundred centimeter of water pressure pulse will produce a  
0.11 atmosphere, or an 11% increase in the total and therefore partial  
pressure of oxygen. Assuming that the sensor is fast enough to track this  
pressure pulse, an un-pressurized reading of 50% oxygen will increase  
to 55.3% if the sensor is subjected to a pressure cycle of 100cm H2O.  
The reading will rise proportionally less for smaller pressures.  
2.3.4 Discrepancy in Readings  
The AX300-I instrument should be used to measure the oxygen  
concentration exiting another oxygen mixing device or life support  
system (i.e., a blender, incubator or anesthesia machine). The  
information obtained from the AX300-I should never be used to adjust a  
life-support system, but should only be used as an indication that the life  
support system or device may require service and/or calibration.  
When a discrepancy in oxygen readings is detected, the oxygen  
analyzers readings should be verified by checking the AX300-I battery  
condition and calibration using 100% oxygen. If the analyzer can be  
calibrated, the unit can be assumed to be in good working order and  
capable of providing readings to specification. If, after reinstalling the  
unit, the discrepancy in oxygen readings persists, the problem is most  
likely elsewhere (i.e., flow blockage, primary device error, etc.). Further  
investigation should be made until the discrepancy in readings is  
resolved. The troubleshooting section of this manual may provide  
additional assistance in locating the problem.  
Note: The MSDS on this material is available upon request  
through the Teledyne Environmental, Health and Safety  
Coordinator. Contact at (626) 934-1592  
2.3.5 Anesthetic Gases  
2.3.5.1 GASES THAT INDUCE READING ERROR  
When using the R17MED sensor in the presence of anesthetic gases  
such as Halothane, the oxygen reading may fall (see Table below). The  
Teledyne Analytical Instruments  
25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
AX300-I  
magnitude of this error will depend upon the level of oxygen and the  
duration of exposure.  
The anesthetic agents listed in the following table (Halothane,  
Enflurane, Isoflurane, Sevoflurane, and Desflurane) were vaporized into  
a stream of 30% oxygen / 70% nitrous oxide, and the resulting drops in  
oxygen level after an exposure of approximately two hours were noted.  
Exposures in excess of two hours may produce slightly greater  
errors. The errors listed are typical for all oxygen sensors such as the  
R17MED. Exposing the sensor to air or gases that do not contain  
anesthetic agents for a period of time equal to or greater than the exposure  
interval will eliminate the reading error in most cases.  
Teledyne Analytical Instruments  
26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Portable Oxygen Analyzer  
Operation  
Table 2-1: Oxygen Reading Error in a Mixture of Anesthetic Gas  
Gas or Vapor Level  
(Balance: Mixture of 30% O2 / 70% N2O, except where noted)  
Gas or Vapor  
Test Level  
Oxygen Reading  
Error  
Helium  
50%, balance O2  
0%  
Nitrous Oxide  
Carbon Dioxide  
Halothane  
80%, balance O2  
0%  
10%, balance O2  
0%  
4%  
5%  
5%  
5%  
15%  
< 1.5% O2 *  
< 1.5% O2 *  
< 1.5% O2 *  
< 1.5% O2 *  
< 1.5% O2 *  
Enflurane  
Isoflurane  
Sevoflurane  
Desflurane  
* Errors are approximate and may vary based on exposure times and  
concentrations.  
These performances meet or exceed the requirements of ISO 7767:  
1997 (E).  
CAUTION:  
THE AX300-I SHOULD NOT BE USED IN THE  
PRESENCE OF FLAMMABLE ANESTHETICS SUCH  
AS DIETHYL ETHER OR CYCLOPROPANE.  
CAUTION: THE AX300-I, OXYGEN SENSOR AND ASSOCIATED  
HARDWARE ARE NON-STERILE DEVICES. DO NOT  
AUTOCLAVE THE INSTRUMENT OR SENSOR, AS  
THIS WILL DAMAGE THE EQUIPMENT.  
2.3.5.2 CARE AFTER USE IN NITROUS OXIDE  
CAUTION:  
THE R17MED SENSOR SHOULD NOT BE LEFT IN  
NITROUS OXIDE MIXTURES ANY LONGER THAN  
ABSOLUTELY NECESSARY.  
After exposure to nitrous oxide mixtures, the sensor should be left  
in 100% oxygen overnight (e.g., left in a breathing circuit that has been  
flushed with pure oxygen). If this is not practical, when using the tee,  
remove the plastic flow diverter and leave the sensor in room air. If the  
Teledyne Analytical Instruments  
27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
AX300-I  
oxygen reading continues to drop after each use in nitrous oxide the  
sensor should be removed from service. If the sensor can no longer be  
calibrated or if there is any sign of electrolyte leakage, the sensor should  
be disposed of in accordance with local regulations and the Material  
Safety Data Sheet (MSDS) available through Teledyne.  
2.3.6 Cleaning  
The AX300-I and R17MED sensor are non-sterile devices.  
The oxygen analyzer, oxygen sensor, and sensor interconnection  
cable may be cleaned by wiping the surfaces with isopropyl alcohol or a  
mild cleaning solution.  
Before placing instrument in operation make sure all surface are  
dry and unit is properly calibrated.  
When cleaning do not allow liquids or moisture to enter the  
instrument or sensor internal cavities. Do not allow the cleaning solution  
to come in contact with the electrical connections. Do not immerse the  
instrument, oxygen sensor or sensor interconnection cable in water or  
any other liquid.  
Do not expose the instrument, oxygen sensor and interconnection  
cable to steam, ethylene oxide, or radiation sterilization.  
2.4 Do’s and Don’ts  
– DO –  
Read all of the directions before using for the first time.  
Calibrate every 8 hours or before every use.  
Visually inspect the sensor for leakage before each use.  
Calibrate using 100% oxygen and check in air.  
Test batteries regularly and replace when battery indicator  
shows low battery (no bars remaining)  
Make sure keys are locked by using the LOCK/UNLOCK  
key feature.  
Keep the unit, sensor and connections dry, or on the dry side  
of the breathing circuit.  
Recalibrate after replacing the batteries.  
Teledyne Analytical Instruments  
28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Portable Oxygen Analyzer  
Operation  
Recalibrate after replacing the sensor.  
Use properly installed alkaline batteries only.  
Make sure the R17MED sensor is properly attached.  
Remove the plastic flow diverter only when using the tee adapter.  
Remove and save the plastic flow diverter when using the  
sensor in non-flowing applications (incubators, tents, etc.)  
Clean the case with isopropyl alcohol or mild detergent only.  
– DON’T –  
Use this analyzer if you suspect any malfunction.  
Use the instrument in the presence of flammable gases.  
Use anything but alkaline batteries.  
Autoclave or freeze the sensor or instrument.  
Open or try to repair a leaking or broken sensor.  
Immerse the unit or sensor in any liquid.  
Pass hot or cold gas mixtures over the sensor.  
Adjust the reading in air after 100% calibration  
Expose the unit to devices that produce high levels of radio,  
short wave, microwave, x-ray, or high frequency  
interference.  
Use cleaning agents or liquids in the cable receptacles or  
around the battery compartment.  
Place the unit itself in a water vapor-saturated environment.  
Expose the LCD to excessive sunlight.  
Expose the unit to a condensing water environment such as a  
mist tent.  
Teledyne Analytical Instruments  
29  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
AX300-I  
BLANK  
Teledyne Analytical Instruments  
30  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Portable Oxygen Analyzer  
Service Manual  
Service Manual  
3.1 General Service Information  
The Teledyne Model AX300-I portable oxygen analyzer is designed to be  
robust yet compact in size. In order to achieve maximum reliability in a  
microprocessor-based instrument, a single PC board is used which relies exclusively  
on surface mount technology. Without access to specialized probes and test  
equipment, troubleshooting and repair of circuit board components are not feasible.  
A factory replacement of the entire PC board is more cost effective than a field  
repair of an individual component.  
With the exception of replacing the sensor or batteries there are no user-  
serviceable components inside the unit. There are no potentiometers or other  
adjustments to be made within this instrument. If a problem arises with either of  
these models that cannot be corrected by recalibration, changing the batteries or  
replacing the sensor as described in this manual, the unit must be sent back to the  
factory for repair or replacement. See Section 3.10 for instructions on obtaining a  
Return Merchandise Authorization (RMA) number before sending a unit back to  
Teledyne for repair.  
3.2 Overall Maintenance  
The AX300-I instrument requires very little maintenance, other than  
calibration, checking and changing the batteries and sensor, and cleaning the  
plastic housing. Occasional cleaning of the plastic surface can be done with  
isopropyl alcohol. Should any part of the instrument malfunction or fail to  
perform, the unit should be removed from service. There are no user-serviceable  
components within the instrument.  
3.3 Battery Maintenance  
DO: Test batteries regularly. (replace immediately when all 5 bars are  
missing).  
DO: Always use alkaline batteries.  
DO: Recalibrate after replacing batteries.  
The AX300-I instrument incorporates a bar graph that continuously displays  
the approximate amount of useful life remaining on the set of installed batteries.  
Teledyne Analytical Instruments  
31  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Service Manual  
AX300-I  
Excessive alarm activation will wear down the battery faster than usual. The  
minimum detectable change in battery voltage corresponds to an increment of  
about 50 hours, meaning that the battery voltage reading may not change for  
several hours at a time.  
If the analyzer is not used for a period of 30 days or more, the batteries  
should be removed prior to storage.  
3.4 Sensor Maintenance  
DO: check the sensor for damage or leaks before use.  
DO: recalibrate after replacing the sensor.  
DON’T: immerse the R17MED sensor in liquid.  
DON’T autoclave the R17MED sensor.  
DON’T: open or try to repair the sensor.  
Before every use, the sensor, cable and connections should be checked.  
Check the sensor for leaks and condensation. Check the cable for splitting or  
cracked insulation. Make sure the connections are tight and dry.  
In the event that the sensor has been damaged, consult the Material Safety  
Data Sheet in the Appendix for handling guidelines.  
3.5 Calibration  
Incorrect readings can often traced to improper calibration. The AX300-I  
should be calibrated before each use and every 8 hours to maintain accuracy. It  
must be calibrated whenever new batteries are installed. Calibration using  
methods other than described below can lead to improper operation and are  
discouraged.  
Whenever new batteries are installed or removed for any reason, the oxygen  
analyzer defaults to the calibration mode with the LCD display flashing 00.0.  
Only the CAL and ON/OFF (I/O) keys are functional at this point.  
To calibrate the instrument refer back to Section 2.1.4.  
.
Teledyne Analytical Instruments  
32  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Portable Oxygen Analyzer  
Service Manual  
3.6 Gas Sampling  
3.6.1 Humidity  
Humidity does not directly affect the accuracy of the sensor's measurement.  
However, when a nebulizer or other device is used to increase moisture levels in  
gas mixtures, the moisture actually dilutes the mixture. This dilution effect  
decreases the oxygen concentration.  
For example, if an 80% oxygen gas mixture is humidified to saturation at  
room temperature, the resulting gas mixture will contain only 77.5% oxygen.  
Your portable oxygen analyzer accurately measures decreases in the oxygen  
concentration due to the dilution effects of moisture added to gas mixtures.  
As with all oxygen sensors, excessive condensation on the sensing surface  
of the R17MED will block the diffusion of oxygen to the sensor, rendering it  
inoperative. We recommend installing the sensor on the dry side of the breathing  
circuit at all times.  
3.6.2 Temperature  
The R17MED oxygen sensor adjusts for ambient temperature changes in the  
range of 0–40°C (32–106°F). Since the thermistor that compensates for these  
changes is located in the rear of the sensor assembly, it is important that gas  
mixtures, flowing over the front of the sensor, be at room temperature. Reading  
errors may occur if hot gases from a heated humidifier are directed past a sensor  
teed into a breathing circuit.  
A small thermal tracking error may be encountered in application areas  
where the entire sensor assembly is placed in the gas mixture to be analyzed  
(e.g., incubators). Holding the sensor in your hand for more than a few minutes  
can also affect the temperature tracking which appears as a slow drift on the  
LCD. No adjustments should be made during this period since this error will be  
eliminated when both the thermistor and sensing electrode have had sufficient  
time to come to thermal equilibrium. This can take up to 2 hours.  
3.6.3 Pressure  
Virtually all gas sensors and analyzers measure the partial pressure, not the  
percentage, of the gas that they sense. The only time that these instruments can  
accurately read percentages is when the total pressure does not vary over time  
between calibrations and use. For this reason it is important to calibrate the  
AX300-I oxygen sensor at regular intervals. It is recommended that the unit be  
calibrated prior to each use or every 8 hours.  
Teledyne Analytical Instruments  
33  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Service Manual  
AX300-I  
When the sensor is connected to a ventilator circuit, the alternating  
“breathing” pressure cycles generated by the ventilator will be sensed as an  
increase in the oxygen percentage (especially if the sensor is fast enough to sense  
the changes, as is the R17MED). In reality, the percentage of oxygen is not  
changing; it is the total pressure that is increasing producing a corresponding  
increase in the partial pressure of oxygen. A hundred centimeter of water  
pressure pulse will produce a 0.11 atmosphere, or an 11% increase in the total  
and therefore partial pressure of oxygen. Assuming that the sensor is fast enough  
to track this pressure pulse, an un-pressurized reading of 50% oxygen will  
increase to 55.3% if the sensor is subjected to a pressure cycle of 100cm H2O.  
The reading will rise proportionally less for smaller pressures.  
3.6.4 Discrepancy in Readings  
The AX300-I instrument should be used to measure the oxygen  
concentration exiting another oxygen mixing device or life support system (i.e., a  
blender, incubator or anesthesia machine). The information obtained from the  
AX300-I should never be used to adjust a life-support system, but should only be  
used as an indication that the life support system or device may require service  
and/or calibration.  
When a discrepancy in oxygen readings is detected, the oxygen analyzers  
readings should be verified by checking the AX300-I battery condition and  
calibration using 100% oxygen. If the analyzer can be calibrated, the unit can be  
assumed to be in good working order and capable of providing readings to  
specification. If, after reinstalling the unit, the discrepancy in oxygen readings  
persists, the problem is most likely elsewhere (i.e., flow blockage, primary  
device error, etc.). Further investigation should be made until the discrepancy in  
readings is resolved. The troubleshooting section of this manual may provide  
additional assistance in locating the problem.  
Note: The MSDS on this material is available upon request through the  
Teledyne Environmental, Health and Safety Coordinator. Contact  
at (626) 934-1592  
3.6.5 Anesthetic Gases  
3.6.5.1 GASES THAT INDUCE READING ERROR  
When using the R17MED sensor in the presence of anesthetic gases such as  
Halothane, the oxygen reading may fall (see Table below). The magnitude of this  
error will depend upon the level of oxygen and the duration of exposure.  
The anesthetic agents listed in the following table (Halothane, Enflurane,  
Isoflurane, Sevoflurane, and Desflurane) were vaporized into a stream of 30%  
Teledyne Analytical Instruments  
34  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Portable Oxygen Analyzer  
Service Manual  
oxygen / 70% nitrous oxide, and the resulting drops in oxygen level after an  
exposure of approximately two hours were noted.  
Exposures in excess of two hours may produce slightly greater errors. The  
errors listed are typical for all oxygen sensors such as the R17MED. Exposing the  
sensor to air or gases that do not contain anesthetic agents for a period of time  
equal to or greater than the exposure interval will eliminate the reading error in  
most cases.  
Table 3-1: Oxygen Reading Error in a Mixture of Anesthetic Gas  
Gas or Vapor Level  
(Balance: Mixture of 30% O2 / 70% N2O, except where noted)  
Gas or Vapor  
Test Level  
Oxygen Reading  
Error  
Helium  
50%, balance O2  
0%  
Nitrous Oxide  
Carbon Dioxide  
Halothane  
80%, balance O2  
0%  
10%, balance O2  
0%  
4%  
5%  
5%  
5%  
15%  
< 1.5% O2 *  
< 1.5% O2 *  
< 1.5% O2 *  
< 1.5% O2 *  
< 1.5% O2 *  
Enflurane  
Isoflurane  
Sevoflurane  
Desflurane  
* Errors are approximate and may vary based on exposure times and  
concentrations.  
These performances meet or exceed the requirements of ISO 7767: 1997 (E).  
CAUTION:  
THE AX300-I SHOULD NOT BE USED IN THE PRESENCE OF  
FLAMMABLE ANESTHETICS SUCH AS DIETHYL ETHER OR  
CYCLOPROPANE.  
CAUTION: THE AX300-I, OXYGEN SENSOR AND ASSOCIATED  
HARDWARE ARE NON-STERILE DEVICES. DO NOT  
AUTOCLAVE THE INSTRUMENT OR SENSOR, AS THIS WILL  
DAMAGE THE EQUIPMENT.  
Teledyne Analytical Instruments  
35  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Service Manual  
AX300-I  
3.6.5.2 CARE AFTER USE IN NITROUS OXIDE  
CAUTION:  
THE R17MED SENSOR SHOULD NOT BE LEFT IN NITROUS  
OXIDE MIXTURES ANY LONGER THAN ABSOLUTELY  
NECESSARY.  
After exposure to nitrous oxide mixtures, the sensor should be left in 100%  
oxygen overnight (e.g., left in a breathing circuit that has been flushed with pure  
oxygen). If this is not practical, when using the tee, remove the plastic flow  
diverter and leave the sensor in room air. If the oxygen reading continues to drop  
after each use in nitrous oxide the sensor should be removed from service. If the  
sensor can no longer be calibrated or if there is any sign of electrolyte leakage,  
the sensor should be disposed of in accordance with local regulations and the  
Material Safety Data Sheet (MSDS).  
Note: The MSDS on this material is available upon request through the  
Teledyne Environmental, Health and Safety Coordinator. Contact  
at (626) 934-1592  
3.7 Troubleshooting  
The AX300-I oxygen analyzer provides a variety of built-in safety features that  
prevent its use when a fault is detected. When a unit displays the message SENSOR  
and sounds the audible and visual alarm continuously, it is an indication of a faulty  
connection between the sensor and the unit or an expired or faulty sensor. To  
determine where the difficulty lies, refer to the following guidelines in Table 3-2.  
Table 3-2 Troubleshooting  
Symptom  
Why  
What To Do  
New sensor responds  
slow or drifts.  
If the sensor is new  
and was just  
A) Wait 1–2 hours and  
recalibrate.  
removed from its  
sealed bag it may  
need to run for  
several hours.  
Sensor will not read  
below 22 % after  
calibration in 100% O2.  
Calibration in 100%  
was invalid or the  
room air is  
contaminated with  
excess oxygen.  
A) Recalibrate using dry gas  
making sure the reading  
stabilizes before making  
any adjustments.  
B) Make sure that at least 6" (30  
cm.) of tubing is attached to  
the exhaust side of the tee  
Teledyne Analytical Instruments  
36  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Portable Oxygen Analyzer  
Service Manual  
adapter to prevent back filling.  
O2 flow rate should not  
exceed 5 l/min.  
C) Oxygen concentration at the  
sensor is significantly higher  
than 21%. Take the  
instrument to a well-  
ventilated area and check  
the reading again.  
E) Try calibrating with a known  
good sensor; if this fails, see  
symptom "Reading drifts  
over 2–3%..."  
The sensor does not  
react to changes in  
oxygen concentration,  
or the readings are  
unstable and drifting.  
Water is condensing  
on the sensing  
surface.  
A) Remove the sensor from tee  
adapter and unscrew the  
plastic flow diverter. Using  
absorbent tissue or cotton  
swab, gently wipe off sensing  
surface inside threaded  
Electrical interference  
is disrupting the  
electronics  
portion of sensor assembly.  
B) Relocate unit away from  
sources of electrical noise  
such as cauterizing equipment  
and two-way radios.  
The display is flashing  
The unit has detected A) Check sensor cable  
a fault in the signal  
from the sensor.  
connections and make sure  
they are completely  
SENSOR  
inserted into the mating  
connector and the capture  
nut is firmly in place.  
Sensor has expired.  
The sensor has been  
exposed to a gas  
containing little or no  
oxygen.  
B) Expose the sensor in 100%  
O2 and check calibration.  
C) The sensor output has  
fallen to a level where it is  
no longer usable. Replace  
sensor.  
The oxygen reading  
fluctuates or appears to the R17MED detects  
be incorrect.  
Like all O2 sensors,  
A) During calibration, make  
sure there is no restrictions  
on exhaust side of sensor. If  
the reading changes with  
flow, the sensor is  
the changes in the  
partial pressure of  
O2.  
pressurized or there may be  
a leak in the system.  
B) If a high degree of accuracy  
is desired, or the  
Teledyne Analytical Instruments  
37  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Service Manual  
AX300-I  
concentration of O2 is in  
excess of 40%, calibration  
with 100% is recommended.  
C) If humidified gas is used to  
ventilate the patient, water  
vapor actually dilutes the  
gas. See Appendix:  
Humidity, Temperature.  
D) If a blender is used, check  
its calibration. See  
Appendix: Discrepancy in  
Readings.  
The unit has stopped  
working and the LCD is equipped with an  
The AX300-I unit is  
A) Disconnect the batteries and  
inspect the contacts for  
displaying  
alphanumeric figures.  
electronic "watch dog,"  
which analyzers the  
circuitry within the unit  
for potential faults and  
renders the unit  
inoperable until the  
condition is corrected.  
(See Watch dog  
corrosion. Reconnect the  
batteries. If the unit functions  
properly, calibrate the unit  
and reset the alarm values.  
B) Try a new set of batteries.  
C) Increase the distance  
between the unit and any  
source of radio frequency  
interference. The sensor  
cable is a prime source of  
pickup as it can act like an  
antenna. Relocate the  
sensor cable and if possible  
change its coiled length to  
“de-tune” its antenna effect.  
Placing the cable in a  
section below)  
Several conditions can  
activate the "watch  
dog." Dropping the  
unit, poor battery  
connections, and radio  
frequency interference  
are the most common  
causes. See the watch  
dog section for  
different position may also  
help.  
additional information  
No display.  
A) Batteries expired.  
A) Check/replace batteries.  
B) Check battery connections.  
C) Calibrate.  
B) Bad battery  
connection.  
Keys inoperable/cannot The LOCK/UNLOCK  
A) Press LOCK/UNLOCK key  
once. LCD will flash  
turn unit off  
key is activated which  
is preventing key  
operation  
indicating keys are active.  
Cannot adjust  
calibration  
Critical settings require A) If display is not flashing  
two keys be pressed in  
a specific order.  
press lock key once to  
activate keys. Press desired  
function followed by the Up  
and Down key.  
LOCK/UNLOCK key is  
active  
Teledyne Analytical Instruments  
38  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Portable Oxygen Analyzer  
Service Manual  
Note: In the event that none of these procedures produce desired results, remove  
the batteries and return the unit to Teledyne for repair.  
3.8 Watchdog  
The AX300-I is equipped with a watchdog circuit that continuously analyzers the  
electronics for proper operation. If the watchdog detects a failure, one of the following  
codes will appear on the LCD.  
The error codes can appear on the LCD when batteries are first installed,  
during normal operation or if the unit is subjected to extreme shock. In some  
cases an additional digit is used in the error code to supply additional  
information. For example, the error code 6 and 7 are followed by another digit  
listed as (N) in Table 3.3. The error code 65 would indicate that a key is stuck  
and this key is the Silence key.  
In addition to supplying visual error codes, the audio device will beep a  
number of times to indicate the general error in case the display is not functional.  
Note: To reset the watchdog error code. Remove one battery for 5  
seconds and replace. If the error persists contact your locale  
representative or Teledyne for assistance.  
Table 3-3 Error Codes  
Error Code  
Audio Beeps  
Error  
Indeterminate  
2
The watchdog timer has timed out indicating  
a serious software error  
30  
50  
3
5
Analog output is different from the expected  
value. May indicate a shorted or over-loaded  
analog output or a failure of the analog to  
digital converter or digital to analog  
converter circuit.  
The ADC circuit failed during POST.  
Teledyne Analytical Instruments  
39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Service Manual  
AX300-I  
6 (5)  
6
A stuck key has been detected. The second  
digit example (5) on the display shows  
which key is stuck:  
(There is no  
indication of  
which key is  
stuck)  
1 - Batt  
2 - Cal  
5 - Silence  
8 - Key Lock  
7(2)  
7
A failure has occurred during the automatic  
calibration of the digital to analog converter  
(DAC) circuit. The second digit example (2)  
shows the type of digital to analog failure.  
(There is no  
indication of  
the type of  
DAC failure)  
0 - Measurement  
1 - High Test  
2 - Low Test  
3 - Offset Cal  
4 - Gain Cal  
3.9 Other Problems with the Instrument  
Most other problems arise from either mechanical damage from the  
instrument falling from a bench or table, or electronic component  
failure. In these units, repair or troubleshooting the PCB or individual  
component on the board is not feasible. It requires specialized test  
equipment and probes not generally available to the public. Under most  
circumstances a replacement of the entire PC Board is recommended.  
The instrument must be returned to the factory for PCB installation.  
Occasionally, depending on the environment of use, keys can  
become stuck or function erratically due to contamination. Use a mild  
non-abrasive cleaner solution to periodically clean the keypad and  
screen. An aerosol jet spray of the type commonly used to clean  
computer keyboards can be used to dislodge dirt and accumulations  
from the keypad.  
The Error Code Table (Table 3-3) includes a description of certain fault  
codes, which are diagnostic of some common (usually electronic) problem  
with the instrument. Some of these codes refer to specific components on the  
PCB that are problematic or have failed. These codes are useful in reporting  
Teledyne Analytical Instruments  
40  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Portable Oxygen Analyzer  
Appendix  
a problem with your instrument to Teledyne Customer Service. If an error  
code is indicated on your analyzer record the number and report it to the  
Customer Service Department at the address below.  
3.10 Return Authorization for Service  
For any service beyond sensor and battery replacement, the  
instrument must be returned to the factory. A return merchandise  
authorization (RMA) number must be obtained from Teledyne  
Analytical Instruments prior to returning an instrument for service. You  
can request a RMA number via email by contacting us at:  
You can also contact us at the address below.  
Customer Service Department  
TELEDYNE Analytical Instruments  
16830 Chestnut Street  
City of Industry, CA 91749-1580 USA  
Phone (626) 934-1500, Fax (626) 961-2538  
Appendix  
A.1 Specifications  
Range: 0-100% oxygen  
Accuracy: +2% of full scale (at constant temperature  
and pressure)  
Response Time: 90% in less than 8 seconds at 25 °C  
Battery Life: Approximately 2000 hr. continuous use in  
a non-alarm condition  
System Power: 3 AA alkaline batteries.  
Sensor Type: Class R17MED  
Teledyne Analytical Instruments  
41  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix  
AX300-I  
Expected Life: 36 months in air. (10 months when  
continuously exposed to 100% oxygen)  
Dimensions: 2.5" W × 1.25" D × 4.5" H (66 mm x 33-  
mm × 111.5 mm)  
Sensor Cable: Retracted: 2 ft / Extended: 10 ft.  
Storage Temp. 10-30 °C (continuous), 5-50 °C  
(Intermittent)  
Operating Temp: 0-40 °C  
Teledyne Analytical Instruments  
42  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Portable Oxygen Analyzer  
Appendix  
A.2 Spare Parts List  
QTY  
PART NO  
DESCRIPTION  
1
C43690-R17MED  
Micro-Fuel Cell R17MED with  
flow diverter P/N A50057  
3
1
1
1
1
1
1
1
1
B99  
B69934  
A268  
B-74543-B  
D-74459  
C-74461  
B-74462  
B-74463  
B-74466  
“AA” size alkaline battery  
Cable assembly  
Tee adapter (22 mm)  
Front panel assembly (AX300-I)  
Back panel  
Battery door  
Battery door lock  
Mounting clip  
Base assembly  
A.3 Optional Accessories  
1
1
1
1
1
1
1
1
A50057  
CP2345  
CP2344  
B647  
A51589  
A51588  
C53790  
A284  
R17MED flow diverter  
Universal Pole Mounting Clamp  
“V” mount Pole Clamp  
“V” mount Wall Adapter  
Sensor adapter cap, female (22 mm)  
Sensor adapter cap, male (22 mm)  
Calibration assembly  
Universal adapter set for pediatric  
circuits (15mm)  
1
1
1
1
A274  
A283  
B-75554  
B-75555  
Tee adapter, autoclavable  
Tee adapter, metal  
0-1 VDC Interface Cable  
RS 232 Interface Cable  
Schematics are available on request.  
____________________  
A minimum charge is applicable to spare parts orders.  
Note: Orders for replacement parts should include the part  
number (if available) and the model and serial number of  
Teledyne Analytical Instruments  
43  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix  
AX300-I  
the instrument for which the parts are intended.  
Orders should be sent to:  
TELEDYNE Analytical Instruments  
16830 Chestnut Street  
City of Industry, CA 91749-1580  
Phone (626) 934-1500, Fax (626) 961-2538  
or your local representative.  
Teledyne Analytical Instruments  
44  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Portable Oxygen Analyzer  
Appendix  
BLANK  
Teledyne Analytical Instruments  
45  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix  
AX300-I  
Index  
accessories, 46  
moisture, 25, 35  
accuracy, 45  
address, 43, 47  
mounting bracket, 15  
nebulizer, 25, 35  
ALARM SILENCE button, 10, 13, 15  
amplification, 13  
analog output, 22  
analysis range, 10, 45  
anesthetic gas, 27, 37  
anode, 12  
nitrous oxide, 28, 38  
optional accessories. See accessories  
output, 12  
output port, 23  
oxygen sensor, 12  
partial pressure, 26, 35  
power, 15  
audio alarm, 13  
bar graph indicator, 20  
batteries, 10, 19  
power requirements, 10, 45  
R17MED, 10, 12, 24, 28, 34, 38  
reading discrepancy, 26, 36  
reading error, 25, 27, 35, 37  
rear panel, 18  
reconfiguring output, 23  
resetting alarms, 10  
response, 45  
battery compartment, 19  
battery installation, 19  
battery status indicator, 10  
breathing circuits, 24  
cable, 16  
calibration, 20, 21, 34  
calibration interval, 26, 36  
cathode, 12  
restriction on selling. See sale  
restriction  
CHECK SENSOR button, 10, 12, 38  
confined volumes, 24  
copyright, ii  
safety information, iv  
sale restriction, ii  
sensor, 10, 12  
current output. See output  
diffusion sampling, 17  
digital output, 23  
dilution effects, 25, 35  
display, 9  
sensor cable, 45  
sensor failure alarm, 10  
sensor life, 45  
sensor maintenance, 34  
sensor mounting, 17  
set up, 15  
diverter, 13, 17, 24, 29, 38  
drift, 26, 35  
signal output, 22  
electrolyte, 12  
error, 27, 28, 37  
error code, 41, 42  
FCC, ii  
signal processing, 13  
signal processor, 12  
spare parts listing, 46  
specifications, 45  
features, 11  
stand, 15  
front panel AX300, 9  
galvanic cell, 12  
standards, 10  
storage, 34  
humidity, 25, 35  
tables listing, viii  
intended use warning, iv  
jumper, 23  
linear, 10, 22  
LOCK/UNLOCK button, 15  
maintenance, 33  
tee adapter, 24  
Teledyne address, 43, 47  
temperature, 25, 35  
temperature tracking, 26, 35  
thermal equilibrium, 26, 35  
Teledyne Analytical Instruments  
46  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Portable Oxygen Analyzer  
Index  
thermal tracking error, 26, 35  
thermistor, 25, 35  
troubleshooting, 38  
type B equipment, i  
ventilator, 26, 36  
warranty, ii  
watchdog circuit, 41  
water vapor, 22  
website address, 43, 47  
Teledyne Analytical Instruments  
47  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Tektronix Welding System 070 A830 51 User Manual
Telex Speaker IBC8 User Manual
TERK Technologies Radio Antenna TOWER User Manual
Toshiba Cordless Telephone FT 8981 User Manual
Toshiba VCR W 704C User Manual
Ultra electronic Printer Opera 100 User Manual
Uniden Cordless Telephone UH 050XR User Manual
ViewSonic Projector VG920 User Manual
Warwick Stereo Amplifier ProTube IV User Manual
Weber Gas Grill 56069 User Manual