Teledyne Marine Instruments 760 PLUS User Manual

TELEDYNE  
HASTINGS  
INSTRUMENTS  
INSTRUCTION MANUAL  
760 PLUS  
VACUUM GAUGE  
I S O 9 0 0 1  
C E R T I F I E D  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table of Contents  
1. GENERAL INFORMATION............................................................................................................................................ 4  
1.1.  
1.2.  
FEATURES .............................................................................................................................................................. 4  
SPECIFICATIONS..................................................................................................................................................... 5  
2.  
3.  
INSTALLATION.......................................................................................................................................................... 6  
2.1.  
2.2.  
2.3.  
RECEIVING INSPECTION ......................................................................................................................................... 6  
PIEZO TRANSDUCER (760S) INSTALLATION ........................................................................................................... 6  
POWER SUPPLY/DISPLAY INSTALLATION............................................................................................................... 7  
2.3.1. Power Supply/Display Cables and Wiring ....................................................................................................... 7  
OPERATING INFORMATION.................................................................................................................................. 8  
3.1.  
GENERAL ............................................................................................................................................................... 8  
PIEZO TRANSDUCER INPUT POWER ........................................................................................................................ 9  
PIEZO TRANSDUCER OUTPUT SIGNAL .................................................................................................................... 9  
POWER SUPPLY/DISPLAY INPUT POWER ................................................................................................................ 9  
THPS-760 POWER SUPPLY PHYSICAL LAYOUT ..................................................................................................... 9  
3.2.  
3.3.  
3.4.  
3.5.  
3.5.1. Front Panel (Refer to page 19, 7.1.1 for physical layout.)............................................................................... 9  
3.5.2. Rear Panel...................................................................................................................................................... 10  
3.6.  
3.7.  
POWER SUPPLY RELAY SET POINT ADJUSTMENT ................................................................................................ 10  
PRESSURE MEASUREMENTS ................................................................................................................................. 11  
4.  
5.  
6.  
7.  
CALIBRATION.......................................................................................................................................................... 12  
4.1.  
4.2.  
HASTINGS 760 PLUS PIEZO TRANSDUCER ZERO AND SPAN ADJUSTMENT ........................................................... 12  
HASTINGS 760S OPTION-5 TRANSDUCER ZERO AND SPAN ADJUSTMENT ............................................................ 12  
MAINTENANCE........................................................................................................................................................ 13  
5.1.  
5.2.  
POWER SUPPLY/DISPLAY AUTHORIZED MAINTENANCE ...................................................................................... 13  
PIEZO TRANSDUCER AUTHORIZED MAINTENANCE .............................................................................................. 13  
WARRANTY .............................................................................................................................................................. 14  
6.1.  
6.2.  
WARRANTY REPAIR POLICY ................................................................................................................................ 14  
NON-WARRANTY REPAIR POLICY ....................................................................................................................... 14  
DIAGRAMS AND DRAWINGS ............................................................................................................................... 15  
7.1.  
FRONT PANEL ...................................................................................................................................................... 15  
REAR PANEL ........................................................................................................................................................ 15  
MOUNTING........................................................................................................................................................... 16  
PANEL MOUNTING ............................................................................................................................................... 16  
7.2.  
7.3.  
7.4.  
Manual: 147-082010_HPM-76D0o-Pwnlulosa-dVfarocmuuWmw-wG.Saoumgaenuals.com. All Manuals Search And Download.  
Page 3 of 16  
1. General Information  
This manual contains technical and general information relating to the installation, operation, and  
calibration of the Hastings 760 Plus vacuum gauge which is comprised of a piezo vacuum transducer  
(760s) and power supply/display.  
The 760s piezoresistive type transducer has one side of its diaphragm evacuated to a low reference  
pressure then sealed to produce an absolute pressure signal output. This output signal is linear-  
proportional to the applied pressure and is independent of gas composition.  
The power supply/display provides 24 VDC excitation power for the 760s transducer and a four digit  
dual setpoint meter relay for monitoring, measurement and control applications.  
For best performance, this instrument should be operated with the appropriate Hastings accessory  
items. Attempting to use a Hastings product with other manufacturers equipment may result in damage  
to all equipment.  
1.1. Features  
Piezo Vacuum Transducer (760s):  
Linear voltage or current output  
Absolute pressure measurement, independent of gas composition  
Zero and span, adjustable  
Compact size, 1.5” x 1.5” x 3”  
0.1 - 1000 Torr range  
Stand alone operation with 15 - 30 VDC input  
All wetted materials 316 stainless steel  
Optional 3-1/2 digit, LCD display  
Power Supply/Display:  
Excitation output for use with Hastings 760S piezo vacuum transducer  
4 1/2-digit high efficiency red LED display  
Two control relays, SPDT, 2 A (max) @ 240 VAC (max)  
Relay set point, set manually from front panel  
120/240 VAC input power (no hardware settings required)  
Front panel (SP1, SP2) annunciators (LED’S)  
1/8 DIN case, panel mounting  
Manual: 147-082010_HPM-76D0o-Pwnlulosa-dVfarocmuuWmw-wG.Saoumgaenuals.com. All Manuals Search And Download.  
Page 4 of 16  
1.2. Specifications  
Piezo Vacuum Transducer (760s):  
Accuracy ......................................................................................................± 0.25% FS* (760 Torr)  
Operating pressure ............................................................................................. 1000 Torr (nominal)  
Maximum over pressure..........................................................................................2300 Torr(45 psi)  
Operating temperature ............................................................................................................. 0-50°C  
Input voltage .....................................................................................................................15-30 VDC  
Input power............................................................................................ 15 mA @ 24 VDC (no load)  
Output impedance................................................................................................... 1k ohm minimum  
Output signal options:  
Option-01 ....................................................................................................... 0-1000 Torr @ 0-5 VDC  
Option-02 Std. .............................................................................................. 0-1000 Torr @ 0-10 VDC  
Option-03......................................................................................................... 0-1000 Torr @ 2-10 mA  
Option-04......................................................................................................... 0-1000 Torr @ 4-20 mA  
Option-05.......................................................................0-1000 Torr @ 5 or 10 VDC w/digital display  
Size........................................................................................................................ 1.5”x 1.5”x 2.875”  
Wetted material(s).................................................................................................... 316 stainless steel  
Power Supply/Display:  
Digital display.......................................................................................... ± 0.05% of reading ± 1 digit  
Temperature coefficient ..................................................................................................< 80 ppm/°C  
Input power.............................................................................. 120 or 240 volts AC, 50 ma, 50/60 hz  
Output power.....................................................................................24 VDC no regulated @ 30 mA  
Input signal......................................................................................................... 0-10 VDC (nominal)  
Relay(s).......................................................................................................... 2A @ 240 VAC form C  
Size............................................................................................................................ 1.9”x 3.8”x 4.6”  
* Accuracy 0.25% FS (calibration based on a 0.15% RDG standard)  
Manual: 147-082010_HPM-76D0o-Pwnlulosa-dVfarocmuuWmw-wG.Saoumgaenuals.com. All Manuals Search And Download.  
Page 5 of 16  
2. Installation  
This section contains the necessary instructions to assist in placing the HPM-760 Plus digital vacuum  
gauge into operation as quickly and easily as possible. Please read the following thoroughly before  
attempting to install or use the instrument.  
2.1. Receiving Inspection  
Carefully unpack the Hastings product(s). Inspect for any obvious signs of damage to the shipment.  
Immediately advise the carrier who delivered the shipment if any damage is suspected. Check each  
component shipped with the packing list. Insure that all parts listed are present. Other equipment or  
accessories will be listed separately.  
2.2. Piezo Transducer (760s) Installation  
Figure 2-1 shows the Hastings piezo vacuum transducer. A standard 1/8” NPT male stem and  
evacuation port is provided for system connection. VCR and KF stem terminations are also available  
When installing transducers with threaded stem, apply an approved vacuum sealant to the stem threads  
of the transducer. Hand thread transducer into vacuum system until joint is finger tight. Tighten  
transducer module completely using a wrench on the hexagonal portion of the stem.  
Do not grip instrument case to tighten; this will result in permanent damage to transducer assembly.  
When the 760s is purchased independently of the power supply/display, a connector is provided so that  
the transducer can be connected as detailed in the figure below.  
TRANSDUCER PINOUT  
PIN1.............................SIGNAL OUTPUT (COM)  
PIN 2..............................SIGNAL OUTPUT (POS)  
PIN 3............................15-30 VDC INPUT (COM)  
PIN 4..............................15-30 VDC INPUT (POS)  
Manual: 147-082010_HPM-76D0o-Pwnlulosa-dVfarocmuuWmw-wG.Saoumgaenuals.com. All Manuals Search And Download.  
Page 6 of 16  
2.3. Power Supply/Display Installation  
Cut a panel opening to the following dimensions; 3.62” (92mm) wide, 1.77” (45mm) tall. Insert the  
power supply/display through the panel cutout and secure in place using mounting hardware supplied.  
2.3.1. Power Supply/Display Cables and Wiring  
Three cables provide all wire connections for the Hastings 760 Plus and are shown below:  
760 Plus Interconnect Cable  
Fig 2-1  
To power supply/display  
To Piezo Transducer (760s)  
Interconnect cable, 4-cond. # 20 A.W.G., 8-ft length standard  
Power Supply/Display Power Cable  
Fig 2-2  
To power supply/display  
AC power cable, 2-cond w/polarized plug., # 18 A.W.G., 8-ft length standard  
Manual: 147-082010_HPM-76D0o-Pwnlulosa-dVfarocmuuWmw-wG.Saoumgaenuals.com. All Manuals Search And Download.  
Page 7 of 16  
3. Operating Information  
This section contains information on operating the PIEZO Resistive Transducer in any of its various  
configurations. Refer to the appropriate section for information on the mode in use.  
3.1. General  
The piezoresistive transducer is shown schematically in figure 3-1. Basically the membrane contains an  
ion implanted resistive element that is part of a Wheatstone bridge circuit shown in figure 3-2. The  
underside of the transducer has been evacuated and sealed to produce a deflection of the membrane the  
magnitude of which is a function of the differential pressure (ΔP) across the membrane. The resistance of  
the membrane element changes with that ΔP and thus changes the bridge output.  
PIEZO RESISTOR  
DIOXIDE  
DEPLETION  
REGION  
SILICON  
CHIP  
V
B
Fig 3-1  
R+ΔR  
R-ΔR  
R-ΔR  
R+ΔR  
V
OUT  
Fig 3-2  
Manual: 147-082010_HPM-76D0o-Pwnlulosa-dVfarocmuuWmw-wG.Saoumgaenuals.com. All Manuals Search And Download.  
Page 8 of 16  
3.2. Piezo Transducer Input Power  
The 760s piezo transducer operates with a DC input from 15 to 30 volts. Use the 4-pin female  
connector supplied and makes necessary connections as outlined. See illustration in Section 2 of this  
manual.  
3.3. Piezo Transducer Output Signal  
The 760s piezo transducer provides either a linear DC voltage output with pressure or a linear DC  
current output with pressure. The type of output is determined by the instrument option specified. A  
brief description of the option, pressure range and output signal is listed below:  
OPTION No  
PRESSURE RANGE  
(Torr)  
SIGNAL OUT  
(see note  
01  
02  
03  
04  
05  
0-1000  
0-1000  
0-1000  
0-1000  
0-1000  
0-5 VDC  
0-10 VDC  
2-10 mA  
4-20 mA  
0-5 or 10 VDC  
w/LCD Digital Display  
NOTE: The 760 Plus power supply/ display or the 760s option-05, provide a direct pressure  
readout in Torr units.  
3.4. Power Supply/Display Input Power  
The 760 Plus power supply/display must be connected to a source of AC voltage, either 120 or 240 V,  
before attempting to program or setup the unit for operation with the Hastings 760S piezo vacuum  
transducer (see Section 2.3 for cable description).  
3.5. THPS-760 Power Supply Physical Layout  
3.5.1. Front Panel (Refer to page 19, 7.1.1 for physical layout.)  
Digital Display. The THPS-760 has a 4 1/2-digit display that displays continuous readout.  
Set Point Adjustment Pots.  
Two Set Point Adjustment Pots are accessible on the panel and are used to set the desired set points.  
(SP1/SP2) Refer to SET POINT ADJUSTMENTS and figure 1 for specific instructions pertaining to  
setting the set points.  
THE KNOBS ON THE SET POINT ADJUSTMENT POTS ARE PERMANENTLY  
INSTALLED AND CAN NOT BE REMOVED. REMOVING THEM COULD CAUSE  
PERMANENT DAMAGE TO THE SET POINT POTS.  
Set Point Pushbuttons.  
There are two set point pushbuttons (SP1 AND SP2) accessible on the front panel. When depressed,  
the setpoint value is displayed on the LED readout  
SET POINT LED’s. The two LED indicators (SP1 and SP2) will illuminate when the input pressure  
is less than desired (SP1/SP2) set point. The LED’s indicate the corresponding relay is activated  
(energized).  
Manual: 147-082010_HPM-76D0o-Pwnlulosa-dVfarocmuuWmw-wG.Saoumgaenuals.com. All Manuals Search And Download.  
Page 9 of 16  
3.5.2. Rear Panel  
Refer to page 19, 7.1.2 for rear view and specific pin outs and connector arrangements.  
AC Input Connector J1  
Universal input power 110vac to 220 VAC. No hardware settings required.  
J3 Input/Output Connector  
Sensor Input (pin 9) Input signal from HPM 760S sensor  
Sensor Ground (pin 10)  
Input ground  
Signal output (pin 11)  
Auxiliary signal output. (0 to +10vdc)  
Signal output return (pin 12) Auxiliary signal return  
24vdc Output (pin 7)  
24vdc Return (pin 8)  
24vdc Auxiliary output  
24vcd Auxiliary return  
SP1 relay contacts (pins 1, 2,3) SP1 relay SPDT contacts.  
(max)2 amps @ 250vac  
SP2 relay contacts (pins 4, 5, 6)) SP2 relay SPDT contacts. (max)2amps @ 250vac  
(Either relay will energize when input pressure is less than desired set point).  
Grounding Considerations  
If the THPS-760 is being used in an area with excessive electromagnetic interference (EMI), the power  
supply may need to be grounded. This is accomplished by attaching one end of a grounding wire to the  
Thumbscrew located at the rear of the THPS 760. The other end should be connected to the 760 sensor  
ground. A shielded cable may be utilized.  
3.6. Power Supply Relay Set Point Adjustment  
The THPS-760 Power Supply has two separate set points with corresponding LED Indicators and  
relay contact outputs. Each Relay has option of either normally open or Normally Closed contacts.  
1. Ensure 120vac/220vac is supplied to HPS-760 Power Supply. (No switching necessary.)  
2. SP1 setting. Depress SP1 set point switch and hold.  
(a) Turn the SP1 Potentiometer until the desired set point is.  
Displayed on digital display.  
(b) Release switch.  
3. SP2 setting. Depress SP2 set point switch and hold.  
(a) Turn the SP2 Set Point Potentiometer until the desired set point is  
Displayed on digital display.  
(b) Release switch.  
4. With both set point push buttons switches released the digital display will indicate input from  
HPM-760 sensor. The SP1/SP2 set point relays will energize when the input is less than the  
SP1/SP2 set point. Depressing both set point switches at the same time will result in an  
erroneous reading, but will not affect the proper operation of the unit.  
Manual: 147-082010_HPM-76D0o-Pwnlulosa-dVfarocmuuWmw-wG.Saoumgaenuals.com. All Manuals Search And Download.  
Page 10 of 16  
Auxiliary Relay Contacts  
The THPS-760 Power Supply has SP1 and SP2 relays for various external switching/monitoring  
applications. Each relay has a single set of Form “C” contacts. (Choice of normally open or normally  
closed contacts). The relays will energize (pick up) when the input is less than the preset set points as  
adjusted in the Set Point Adjustment procedures above. The set point LED Indicators (SP1 or SP2) will  
illuminate indicating that (SP1 or SP2) relay is energized. The choice of the normally open or normally  
closed contacts will depend on the desired switching function of the particular application. Each set of  
contacts is capable of switching up to 2 amps (max) @ 250 vac .  
3.7. Pressure Measurements  
The 760 Plus piezo transducers are shipped from the factory fully calibrated; thus, can provide  
immediate and accurate pressure measurements as soon as the installation outlined in Section 2.2 has  
been completed. If using a 760 Plus power supply/display, the pressure can be read directly in Torr units.  
When not using a 760 Plus power supply/display, the linear output voltage or current depending on  
option, must be interpolated to obtain absolute pressure measurements.  
Figure 3-3 on the following page of this manual shows a typical signal output vs pressure curve. Note  
the excellent linearity over 4-decades of pressure.  
The Hastings 760 Plus power supply/display is calibrated at the factory prior to shipment and does not  
require adjustment when installed. The user should not attempt to calibrate the Hastings 760 Plus power  
supply/display; however, if the display indication is incorrect, calibrate the Hastings piezo transducer by  
adjusting the ZERO and SPAN controls as outlined in this manual.  
The Hastings transducer can be adjusted by performing the following procedure. If you intend to use  
this procedure, you will need a vacuum system capable of achieving base pressures below 1.00E-02 Torr,  
and a power supply/display which is calibrated for the option in use.  
1000.0  
100.0  
5.0  
0.5  
INDICATED  
PRESSURE  
(TORR)  
OUTPUT  
VOLTS  
10.0  
0.05  
1.0  
0.005  
0.0005  
0.1  
0.10  
1.00  
10.00  
100.00  
1000.00  
PRESSURE (TORR)  
Manual: 147-082010_HPM-76D0o-Pwnlulosa-dVfarocmuuWmw-wG.Saoumgaenuals.com. All Manuals Search And Download.  
Page 11 of 16  
4.Calibration  
4.1. Hastings 760 Plus Piezo Transducer Zero and Span Adjustment  
1. Install the transducer and power supply/display then apply operating power. Wait a mini  
mum of 15 minutes before proceeding.  
2. Evacuate system to a pressure less than 0.010 Torr (1.00E-02 Torr).  
3. Adjust transducer zero trimpot (Fig. 2-1) until the power supply/display  
indicates 000.0 ±000.1 Torr.  
4. Increase system pressure to atmosphere and adjust span trimpot until the power supply/  
display indicates the known local atmospheric pressure ±000.1 Torr.  
5. Repeat steps 2 through 4 if necessary.  
4.2. Hastings 760s Option-5 Transducer Zero and Span Adjustment  
1. Install the transducer and apply operating power. Wait a minimum of 15 minutes before  
proceeding.  
2. Evacuate the system to a pressure less than 0.01 Torr (1.00E-02 Torr).  
3. Adjust transducer zero trimpot (Fig. 2-1) until display indicates 000 ± 001 Torr.  
4. Increase system pressure to atmosphere and adjust span trimpot (Fig. 2-1) until display  
indicates the known local atmospheric pressure ±001 Torr.  
5. Repeat steps 2 through 4 if necessary.  
Manual: 147-082010_HPM-76D0o-Pwnlulosa-dVfarocmuuWmw-wG.Saoumgaenuals.com. All Manuals Search And Download.  
Page 12 of 16  
5.Maintenance  
5.1. Power Supply/Display Authorized Maintenance  
During normal operation and excluding the calibration procedure, Hastings 760 Plus power  
supply/display does not require maintenance. In the event a Hastings power supply/display fails or is  
otherwise damaged, it should be returned to the manufacturer (factory) for repair or replacement after  
first obtaining the necessary returned material authorization (RMA) number.  
5.2. Piezo Transducer Authorized Maintenance  
During normal operation and excluding the calibration procedure, Hastings piezo vacuum transducers  
(760S) do not require maintenance or adjustment. In the event a transducer fails or is otherwise  
damaged, it should be returned to the manufacturer (factory) for repair or replacement after first  
obtaining the necessary returned material authorization (RMA) number.  
Manual: 147-082010_HPM-76D0o-Pwnlulosa-dVfarocmuuWmw-wG.Saoumgaenuals.com. All Manuals Search And Download.  
Page 13 of 16  
6. WARRANTY  
6.1. Warranty Repair Policy  
Hastings Instruments warrants this product for a period of one year from the date of shipment to be  
free from defects in material and workmanship. This warranty does not apply to defects or failures  
resulting from unauthorized modification, misuse or mishandling of the product. This warranty does not  
apply to batteries or other expendable parts, or to damage caused by leaking batteries or any similar  
occurrence. This warranty does not apply to any instrument which has had a tamper seal removed or  
broken.  
This warranty is in lieu of all other warranties, expressed or implied, including any implied warranty as  
to fitness for a particular use. Hastings Instruments shall not be liable for any indirect or consequential  
damages.  
Hastings Instruments, will, at its option, repair, replace or refund the selling price of the product if  
Hastings Instruments determines, in good faith, that it is defective in materials or workmanship during the  
warranty period. Defective instruments should be returned to Hastings Instruments, shipment prepaid,  
together with a written statement of the problem and a Return Material Authorization (RMA) number.  
Please consult the factory for your RMA number before returning any product for repair. Collect  
freight will not be accepted.  
6.2. Non-Warranty Repair Policy  
Any product returned for a non-warranty repair must be accompanied by a purchase order, RMA form  
and a written description of the problem with the instrument. If the repair cost is higher, you will be  
contacted for authorization before we proceed with any repairs. If you then choose not to have the  
product repaired, a minimum will be charged to cover the processing and inspection. Please consult the  
factory for your RMA number before returning any product repair.  
TELEDYNE HASTINGS INSTRUMENTS  
804 NEWCOMBE AVENUE  
HAMPTON, VIRGINIA 23669 U.S.A.  
ATTENTION: REPAIR DEPARTMENT  
TELEPHONE  
TOLL FREE  
FAX  
(757) 723-6531  
1-800-950-2468  
(757) 723-3925  
E MAIL  
INTERNET  
Repair Forms may be obtained from the “Information Request” section of the Hastings Instruments  
web site.  
Manual: 147-082010_HPM-76D0o-Pwnlulosa-dVfarocmuuWmw-wG.Saoumgaenuals.com. All Manuals Search And Download.  
Page 14 of 16  
7. Diagrams and Drawings  
7.1. Front Panel  
The THPS-760 has a 4 1/2-digit display that displays continuous readout.  
7.2. Rear Panel  
Decal is to be black ink printed on silver compucal, with adhesive on the back  
Manual: 147-082010_HPM-76D0o-Pwnlulosa-dVfarocmuuWmw-wG.Saoumgaenuals.com. All Manuals Search And Download.  
Page 15 of 16  
7.3. Mounting  
Cut out panel as shown in the diagram below.  
7.4. Panel Mounting  
1) Slide power supply into panel from the top.  
2) Attach brackets to the side of the power supply.  
3) Tighten screws against panel.  
Manual: 147-082010_HPM-76D0o-Pwnlulosa-dVfarocmuuWmw-wG.Saoumgaenuals.com. All Manuals Search And Download.  
Page 16 of 16  

Telular Network Router SX3i User Manual
Therma Tru Door Light Commercial Multi Family 22 24 Gauge Steel Edged Door User Manual
Toastmaster Electric Steamer EST7CAN User Manual
Topcom Blood Pressure Monitor BPM WRIST 2301B User Manual
Toshiba Power Supply QX7 Series User Manual
Toyota Automobile Parts P1335 User Manual
Tripp Lite TV Cables P570 006 User Manual
Tyan Computer Network Card B5381 User Manual
Vector Automobile Parts VEC028AC User Manual
VTech Cordless Telephone ip8301 User Manual