Teledyne Doll 200AU User Manual

INSTRUCTION MANUAL  
MODEL 200AU  
NITROGEN OXIDES ANALYZER  
TELEDYNE INSTRUMENTS  
ADVANCED POLLUTION INSTRUMENTATION DIVISION  
(T-API)  
6565 NANCY RIDGE DRIVE  
SAN DIEGO, CA 92121-2251  
TOLL-FREE: 800-324-5190  
FAX: 858-657-9816  
TEL: 858-657-9800  
WEB SITE: www.teledyne-api.com  
02293  
REV. F  
Copyright 1999 API Inc.  
07/06/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
TABLE OF CONTENTS  
SAFETY MESSAGES .........................................................................................II  
TABLE OF CONTENTS.....................................................................................III  
LIST OF FIGURES........................................................................................... VII  
LIST OF TABLES............................................................................................ VIII  
1 HOW TO USE THIS MANUAL..................................................................... 1-1  
2 GETTING STARTED.................................................................................... 2-1  
2.1 UNPACKING....................................................................................................................2-1  
2.2 ELECTRICAL AND PNEUMATIC CONNECTIONS ....................................................................2-1  
2.3 INITIAL OPERATION .........................................................................................................2-6  
3 SPECIFICATIONS, AGENCY APPROVALS, WARRANTY ........................ 3-1  
3.1 SPECIFICATIONS.............................................................................................................3-1  
3.2 EPA EQUIVALENCY DESIGNATION ...................................................................................3-2  
3.3 WARRANTY ....................................................................................................................3-3  
4 THE M200AU NOX ANALYZER................................................................... 4-1  
4.1 PRINCIPLE OF OPERATION...............................................................................................4-1  
4.2 OPERATION SUMMARY ....................................................................................................4-3  
4.2.1 Sensor Module, Reaction Cell, Detector................................................................4-3  
4.2.2 Pneumatic Sensor Board.......................................................................................4-3  
4.2.3 Computer Hardware and Software ........................................................................4-4  
4.2.4 V/F Board ..............................................................................................................4-4  
4.2.5 Front Panel............................................................................................................4-5  
4.2.6 Power Supply Module............................................................................................4-7  
4.2.7 Pump, Valves, Pneumatic System.........................................................................4-7  
4.2.8 Ozone Generator...................................................................................................4-7  
4.2.9 Molybdenum Converter – Ozone Scrubber ...........................................................4-8  
5 SOFTWARE FEATURES............................................................................. 5-1  
5.1 INDEX TO FRONT PANEL MENUS......................................................................................5-1  
5.1.1 Sample Menu ........................................................................................................5-4  
5.1.2 Set-Up Menu .........................................................................................................5-6  
iii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
5.2 SAMPLE MODE .............................................................................................................5-10  
5.2.1 Test Functions.....................................................................................................5-10  
5.2.2 CAL, CALS, CALZ, Calibration Functions............................................................5-13  
5.3 SET-UP MODE .............................................................................................................5-16  
5.3.1 Configuration Information (CFG) .........................................................................5-16  
5.3.2 Automatic Calibration (AutoCal) ..........................................................................5-16  
5.3.3 Data Acquisition System (DAS)...........................................................................5-16  
5.3.4 Range Menu........................................................................................................5-19  
5.3.5 Password Enable.................................................................................................5-21  
5.3.6 Time of Day Clock ...............................................................................................5-22  
5.3.7 Diagnostic Mode..................................................................................................5-22  
5.3.8 Communications Menu........................................................................................5-22  
5.3.9 Variables Menu (VARS) ......................................................................................5-22  
5.3.10 M200AU Operating Modes................................................................................5-22  
5.4 STATUS OUTPUT ..........................................................................................................5-23  
5.5 RS-232 INTERFACE......................................................................................................5-24  
5.5.1 Setting Up the RS-232 Interface..........................................................................5-25  
5.5.2 Command Summary............................................................................................5-28  
5.5.3 TEST Commands and Messages........................................................................5-32  
5.5.4 WARNING Commands and Messages................................................................5-33  
5.5.5 CALIBRATION Commands and Messages.........................................................5-34  
5.5.6 DIAGNOSTIC Commands and Messages...........................................................5-36  
5.5.7 DAS Commands and Message ...........................................................................5-37  
5.5.8 Internal Variables.................................................................................................5-39  
6 OPTIONAL HARDWARE AND SOFTWARE............................................... 6-1  
6.1 RACK MOUNT OPTIONS...................................................................................................6-1  
6.2 ZERO/SPAN VALVES .......................................................................................................6-1  
6.3 AUTOCAL - SETUP ZERO/SPAN VALVES............................................................................6-2  
6.4 4-20 MA, CURRENT LOOP OUTPUT..................................................................................6-4  
6.5 NOY CONVERTER ...........................................................................................................6-4  
7 CALIBRATION & ZERO/SPAN CHECKS ................................................... 7-1  
7.1 MANUAL ZERO/SPAN CHECK OR CAL WITH ZERO/SPAN GAS IN THE SAMPLE PORT..............7-4  
7.2 MANUAL ZERO/SPAN CHECK OR CALIBRATION WITH ZERO/SPAN VALVES OPTION ...............7-6  
7.3 AUTOMATIC ZERO/SPAN CHECK ......................................................................................7-7  
7.4 DYNAMIC ZERO/SPAN CALIBRATION .................................................................................7-7  
7.5 USE OF ZERO/SPAN VALVES WITH REMOTE CONTACT CLOSURE ........................................7-8  
7.6 EPA PROTOCOL CALIBRATION ........................................................................................7-9  
7.6.1 Calibration of Equipment .......................................................................................7-9  
7.6.2 Calibration Gas and Zero Air Sources.................................................................7-11  
7.6.3 Data Recording Device........................................................................................7-12  
7.6.4 Gas Phase Titration (GPT) System .....................................................................7-12  
iv  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
7.6.5 Dynamic Multipoint Calibration Procedure...........................................................7-16  
7.6.6 Moly Converter Efficiency....................................................................................7-23  
7.6.7 Calibration Frequency .........................................................................................7-25  
7.6.8 Other Quality Assurance Procedures ..................................................................7-25  
7.6.9 Summary of Quality Assurance Checks ..............................................................7-27  
7.6.10 ZERO and SPAN Checks..................................................................................7-29  
7.6.11 Recommended Standards for Establishing Traceability ....................................7-30  
7.6.12 Certification Procedures of Working Standards.................................................7-31  
7.7 CALIBRATION OF INDEPENDENT RANGES OR AUTORANGING .............................................7-32  
7.7.1 Zero Calibration with AutoRange or Independent Range ....................................7-32  
7.7.2 Span Calibration with AutoRange or Independent Range ...................................7-33  
7.8 CALIBRATION QUALITY ..................................................................................................7-33  
7.9 REFERENCES ...............................................................................................................7-35  
8 MAINTENANCE ........................................................................................... 8-1  
8.1 MAINTENANCE SCHEDULE ...............................................................................................8-1  
8.2 REPLACING THE SAMPLE PARTICULATE FILTER .................................................................8-3  
8.3 SAMPLE PUMP MAINTENANCE..........................................................................................8-5  
8.4 CLEANING THE REACTION CELL .......................................................................................8-7  
8.5 REPLACING THE MOLYBDENUM CONVERTER .....................................................................8-9  
8.6 PNEUMATIC LINE INSPECTION ........................................................................................8-11  
8.7 LEAK CHECK PROCEDURE.............................................................................................8-15  
8.8 LIGHT LEAK CHECK PROCEDURE ...................................................................................8-16  
8.9 PROM REPLACEMENT PROCEDURE ................................................................................8-16  
9 TROUBLESHOOTING, ADJUSTMENTS .................................................... 9-1  
9.1 OPERATION VERIFICATION-M200AU DIAGNOSTIC TECHNIQUES.........................................9-3  
9.1.1 Fault Diagnosis with TEST Variables ....................................................................9-3  
9.1.2 Fault Diagnosis with WARNING Messages...........................................................9-8  
9.1.3 Fault Diagnosis using DIAGNOSTIC Mode .........................................................9-11  
9.1.4 M200AU Internal Variables..................................................................................9-19  
9.1.5 Test Channel Analog Output ...............................................................................9-21  
9.1.6 Factory Calibration Procedure.............................................................................9-23  
9.2 PERFORMANCE PROBLEMS............................................................................................9-26  
9.2.1 AC Power Check .................................................................................................9-26  
9.2.2 Flow Check..........................................................................................................9-27  
9.2.3 No Response to Sample Gas ..............................................................................9-27  
9.2.4 Negative Output...................................................................................................9-28  
9.2.5 Excessive Noise ..................................................................................................9-28  
9.2.6 Unstable Span.....................................................................................................9-29  
9.2.7 Unstable Zero......................................................................................................9-30  
9.2.8 Inability to Span...................................................................................................9-30  
9.2.9 Inability to Zero....................................................................................................9-31  
v
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
9.2.10 Non-Linear Response........................................................................................9-31  
9.2.11 Slow Response..................................................................................................9-32  
9.2.12 Analog Output Doesn't Agree with Display Concentration.................................9-32  
9.3 SUBSYSTEM TROUBLESHOOTING AND ADJUSTMENTS.......................................................9-32  
9.3.1 Computer, Display, Keyboard..............................................................................9-32  
9.3.2 RS-232 Communications.....................................................................................9-36  
9.3.3 Voltage/Frequency (V/F) Board...........................................................................9-39  
9.3.4 Status/Temp Board..............................................................................................9-45  
9.3.5 Power Supply Module..........................................................................................9-47  
9.3.6 Ozone Generator.................................................................................................9-52  
9.3.7 Flow/Pressure Sensor .........................................................................................9-55  
9.3.8 NOx Sensor Module.............................................................................................9-60  
9.3.9 Z/S Valves...........................................................................................................9-64  
9.3.10 Pneumatic System.............................................................................................9-65  
10 M200AU SPARE PARTS LIST ................................................................ 10-1  
APPENDIX A ELECTRICAL SCHEMATICS ..................................................A-1  
vi  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
LIST OF FIGURES  
FIGURE 2-1: REMOVAL OF SHIPPING SCREWS & CHECK FOR CORRECT POWER ..........................2-3  
FIGURE 2-2: REAR PANEL........................................................................................................2-4  
FIGURE 2-3: INLET AND EXHAUST VENTING ...............................................................................2-5  
FIGURE 2-4: FRONT PANEL....................................................................................................2-10  
FIGURE 2-5: ASSEMBLY LAYOUT ............................................................................................2-11  
FIGURE 4-1: BLOCK DIAGRAM..................................................................................................4-2  
FIGURE 5-1: SAMPLE MENU TREE ............................................................................................5-2  
FIGURE 5-2: SETUP MENU TREE ..............................................................................................5-3  
FIGURE 7-1: CALIBRATION SETUP ............................................................................................7-3  
FIGURE 7-2: DIAGRAM OF GPT CALIBRATION SYSTEM.............................................................7-18  
FIGURE 8-1: REPLACING THE PARTICULATE FILTER ...................................................................8-4  
FIGURE 8-2: SAMPLE PUMP ASSEMBLY.....................................................................................8-6  
FIGURE 8-3: REACTION CELL ASSEMBLY ..................................................................................8-8  
FIGURE 8-4: MOLYBDENUM CONVERTER ASSEMBLY ................................................................8-10  
FIGURE 8-5: PNEUMATIC DIAGRAM - STANDARD CONFIGURATION.............................................8-12  
FIGURE 8-6: PNEUMATIC DIAGRAM WITH ZERO/SPAN VALVE OPTION ........................................8-13  
FIGURE 8-7: PNEUMATIC DIAGRAM WITH EXTERNAL CONVERTER OPTION .................................8-14  
FIGURE 9-1: SPAN CALIBRATION VOLTAGE .............................................................................9-25  
FIGURE 9-2: CPU BOARD JUMPER SETTINGS .........................................................................9-34  
FIGURE 9-3: RS-232 PIN ASSIGNMENTS ...............................................................................9-38  
FIGURE 9-4: V/F BOARD SETTINGS........................................................................................9-44  
FIGURE 9-5: POWER SUPPLY MODULE LAYOUT.......................................................................9-49  
FIGURE 9-6: ELECTRICAL BLOCK DIAGRAM .............................................................................9-50  
FIGURE 9-7: OZONE GENERATOR SUBSYSTEM........................................................................9-54  
FIGURE 9-8: FLOW/PRESSURE SENSOR..................................................................................9-57  
FIGURE 9-9: NOX SENSOR MODULE .......................................................................................9-58  
FIGURE 9-10: NOX SENSOR MODULE .....................................................................................9-59  
FIGURE 9-11: PMT COOLER SUBSYSTEM...............................................................................9-61  
FIGURE 9-12: HIGH VOLTAGE POWER SUPPLY........................................................................9-63  
vii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
LIST OF TABLES  
TABLE 2-1: FINAL TEST AND CALIBRATION VALUES..................................................................2-12  
TABLE 2-1: FINAL TEST AND CALIBRATION VALUES (CONTINUED) .............................................2-13  
TABLE 4-1: FRONT PANEL STATUS LED'S.................................................................................4-6  
TABLE 4-2: OZONE GENERATOR START-UP TIMING....................................................................4-8  
TABLE 5-1: SAMPLE MENU ......................................................................................................5-4  
TABLE 5-2: SAMPLE MENU ......................................................................................................5-5  
TABLE 5-3: SETUP MENU.........................................................................................................5-6  
TABLE 5-3: SETUP MENU (CONTINUED)....................................................................................5-7  
TABLE 5-4: SETUP MENU.........................................................................................................5-8  
TABLE 5-5: SETUP MENU.........................................................................................................5-9  
TABLE 5-6: DAS DATA CHANNEL EDITING ..............................................................................5-19  
TABLE 5-7: CALIBRATE, SETUP PASSWORDS ..........................................................................5-21  
TABLE 5-8: M200AU OPERATING MODES ..............................................................................5-23  
TABLE 5-9: STATUS OUTPUT PIN ASSIGNMENTS......................................................................5-24  
TABLE 5-10: RS-232 PORT SETUP - FRONT PANEL ................................................................5-25  
TABLE 5-11: RS-232 SWITCHING FROM TERMINAL MODE TO COMPUTER MODE .......................5-27  
TABLE 5-12: RS-232 TERMINAL MODE EDITING KEYS.............................................................5-27  
TABLE 5-13: RS-232 COMMAND SUMMARY............................................................................5-29  
TABLE 5-14: RS-232 COMMAND SUMMARY............................................................................5-30  
TABLE 5-15: RS-232 INTERFACE COMMAND TYPES ................................................................5-31  
TABLE 5-16: RS-232 TEST MESSAGES ..................................................................................5-32  
TABLE 5-17: RS-232 WARNING MESSAGES ...........................................................................5-33  
TABLE 5-18: RS-232 CALIBRATION COMMANDS......................................................................5-34  
TABLE 5-19: RS-232 CALIBRATION EXAMPLES .......................................................................5-35  
TABLE 5-20: RS-232 CALIBRATION MESSAGES ......................................................................5-36  
TABLE 5-21: RS-232 DIAGNOSTIC COMMAND SUMMARY .........................................................5-37  
TABLE 6-1: ZERO/SPAN VALVE OPERATION ..............................................................................6-1  
TABLE 6-2: EXAMPLE OF AUTOCAL SETUP................................................................................6-4  
TABLE 7-1: TYPES OF ZERO/SPAN CHECK AND CALIBRATION .....................................................7-2  
TABLE 7-2: MANUAL ZERO CALIBRATION PROCEDURE - ZERO GAS THRU SAMPLE PORT ..............7-4  
TABLE 7-3: ENTER EXPECTED SPAN GAS CONCENTRATIONS PROCEDURE ..................................7-5  
TABLE 7-4: MANUAL SPAN CALIBRATION PROCEDURE - SPAN GAS THRU SAMPLE PORT ..............7-5  
TABLE 7-5: MANUAL ZERO CALIBRATION PROCEDURE - Z/S VALVES ..........................................7-6  
TABLE 7-6: MANUAL SPAN CALIBRATION PROCEDURE - Z/S VALVES ..........................................7-7  
TABLE 7-7: ENABLING DYNAMIC ZERO/SPAN.............................................................................7-8  
TABLE 7-8: Z/S VALVES MODES WITH REMOTE CONTACT CLOSURE ...........................................7-9  
TABLE 7-9: ACTIVITY MATRIX FOR CALIBRATION EQUIPMENT AND SUPPLIES ..............................7-10  
TABLE 7-10: ACTIVITY MATRIX FOR CALIBRATION PROCEDURE.................................................7-11  
TABLE 7-11: ZERO CALIBRATION PROCEDURE ........................................................................7-19  
TABLE 7-12: EXPECTED SPAN GAS CONCENTRATION PROCEDURE...........................................7-20  
viii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
TABLE 7-13: SPAN CALIBRATION PROCEDURE.........................................................................7-20  
TABLE 7-14: AUTOMATIC CALCULATION OF CONVERTER EFFICIENCY ........................................7-24  
TABLE 7-15: DEFINITION OF LEVEL 1 AND LEVEL 2 ZERO AND SPAN CHECKS ............................7-26  
TABLE 7-16: ACTIVITY MATRIX FOR DATA QUALITY..................................................................7-28  
TABLE 7-17: NIST-SRM'S AVAILABLE FOR TRACEABILITY OF CALIBRATION AND AUDIT GAS STANDARDS  
........................................................................................................................7-31  
TABLE 7-18: CALIBRATION QUALITY CHECK ............................................................................7-34  
TABLE 8-1: PREVENTATIVE MAINTENANCE SCHEDULE................................................................8-1  
TABLE 8-2: PREVENTATIVE MAINTENANCE CALENDAR...............................................................8-2  
TABLE 9-1: TEST FUNCTIONS...................................................................................................9-4  
TABLE 9-1: TEST FUNCTIONS (CONTINUED) ..............................................................................9-5  
TABLE 9-1: TEST FUNCTIONS (CONTINUED) ..............................................................................9-6  
TABLE 9-1: TEST FUNCTIONS (CONTINUED) ..............................................................................9-7  
TABLE 9-2: FRONT PANEL WARNING MESSAGES .......................................................................9-9  
TABLE 9-2: FRONT PANEL WARNING MESSAGES (CONTINUED).................................................9-10  
TABLE 9-3: SUMMARY OF DIAGNOSTIC MODES........................................................................9-12  
TABLE 9-4: DIAGNOSTIC MODE - SIGNAL I/O...........................................................................9-13  
TABLE 9-4: DIAGNOSTIC MODE - SIGNAL I/O (CONTINUED)......................................................9-14  
TABLE 9-4: DIAGNOSTIC MODE - SIGNAL I/O (CONTINUED)......................................................9-15  
TABLE 9-4: DIAGNOSTIC MODE - SIGNAL I/O (CONTINUED)......................................................9-16  
TABLE 9-5: MODEL 200AU VARIABLES...................................................................................9-20  
TABLE 9-6: TEST CHANNEL READINGS....................................................................................9-21  
TABLE 9-6: TEST CHANNEL READINGS (CONTINUED) ...............................................................9-22  
TABLE 9-7: MOTHERBOARD JUMPER SETTINGS .......................................................................9-41  
TABLE 9-8: V/F BOARD SWITCH SETTINGS .............................................................................9-42  
TABLE 9-9: POWER SUPPLY MODULE SUBASSEMBLIES ............................................................9-48  
TABLE 9-10: POWER SUPPLY MODULE LED OPERATION .........................................................9-51  
TABLE 9-11: OZONE GENERATOR CONTROL CONDITIONS ........................................................9-53  
TABLE 10-1: TELEDYNE API M200AU SPARE PARTS LIST...................................................10-1  
TABLE 10-1: TELEDYNE API M200AU SPARE PARTS LIST (CONTINUED) ..............................10-2  
TABLE 10-2: TELEDYNE API MODEL 200AU EXPENDABLE KIT ...........................................10-3  
TABLE A-1: ELECTRICAL SCHEMATICS ..................................................................................... A-1  
ix  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
INTENTIONALLY BLANK  
x
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
1 HOW TO USE THIS MANUAL  
The Model 200AU has been designed to provide serviceability, reliability and ease of operation.  
The M200AU’s microprocessor continually checks operating parameters such as temperature,  
flow, and critical voltages. The instrument’s modular design uses captive screws to facilitate  
repair and ease of access. If you encounter any difficulty refer to Section 9 General  
Troubleshooting Hints.  
We recognize that the need for information in this manual changes as time passes. When the  
instrument first arrives, it is necessary to get it up and running quickly and verify its correct  
operation. As time passes, more detailed information is often required on special configurations,  
calibration alternatives and other operational details. Finally there is the need for periodic  
maintenance and to quickly troubleshoot problems to assure maximum reliability and data  
integrity.  
To address these needs, we have created three indexes to the information inside. They are:  
Table of Contents:  
Outlines the contents of the manual in the order the information is presented. This is a good  
overview of the topics covered in the manual. There is also a list of Tables and a list of Figures.  
Index to M200AU Front Panel Menus:  
The Menu Index (Figure 5-1, Figure 5-2, Table 5-1 and Table 5-2) briefly describes the front  
panel menus and refers you to other sections of the manual that have a detailed explanation of  
each menu selection.  
Troubleshooting Section 9:  
The Troubleshooting Section, outlined in the Table of Contents, allows you to diagnose and  
repair the instrument based on variables in the TEST menu, the results of DIAGNOSTIC tests,  
and performance faults such as excessive noise or drift. The troubleshooting section also  
explains the operation, adjustment, diagnosis and testing of each instrument subsystem.  
If you are unpacking the instrument for the first time, please refer to Getting Started in  
Section 2.  
1-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
INTENTIONALLY BLANK  
1-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
2 GETTING STARTED  
2.1 Unpacking  
1.  
Verify that there is no apparent shipping damage. If damage has occurred please advise  
shipper first, then Teledyne API.  
CAUTION  
To avoid personal injury, always use two persons to  
lift and carry the Model 200AU.  
2.  
3.  
4.  
Before operation, it is necessary to remove the shipping hold-down screws. Remove the  
instrument cover, then refer to Figure 2-1 for screw location.  
Also check for internal shipping damage, and generally inspect the interior of the  
instrument to make sure all circuit boards and other components are in good shape.  
Please check the voltage and frequency label on the rear panel of the instrument for  
compatibility with the local power before plugging in the M200AU.  
2.2 Electrical and Pneumatic Connections  
Refer to Figure 2-2 to locate the rear panel electrical and pneumatic connections.  
1.  
2.  
Attach the pump to the Exhaust Out port on the instrument rear panel.  
If you are connecting to a calibrator, attach a vented sample inlet line to the sample inlet  
port. The pressure of the sample gas at the inlet port should be at ambient pressure. The  
exhaust from the pump should be vented to atmospheric pressure. See Figure 2-3 for inlet  
and exhaust line venting recommendations during calibration.  
3.  
4.  
If desired, attach the analog output connections to a strip chart recorder and/or  
datalogger. Refer to Figure 9-4 for the jumper settings for the desired analog output voltage  
range. Factory default setting is 0-5VDC.  
Connect the power cord to the correct voltage line, then turn to Section 2.3 Initial  
Operation.  
2-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
WARNING  
Analyzer Exhaust –Sample Pump  
For brief periods after power-up, analyzer  
exhaust may contain ozone.  
Make sure pump exhaust is routed to well ventilated  
area at atmospheric pressure.  
WARNING  
Lethal voltages present inside case.  
Do not operate with cover off during normal operation.  
Before operation check for correct input voltage and frequency for  
both the analyzer and the sample pump.  
Do not operate without proper chassis grounding.  
Do not defeat the ground wire on power plug.  
Turn off analyzer power before disconnecting  
electrical subassemblies.  
2-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Figure 2-1: Removal of Shipping Screws & Check for Correct Power  
2-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Figure 2-2: Rear Panel  
2-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Figure 2-3: Inlet and Exhaust Venting  
2-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
2.3 Initial Operation  
1.  
2.  
After confirming proper supply voltage, turn on the instrument power.  
The display should immediately light, displaying the instrument type (M200AU) and the  
computer's memory configuration. If you are unfamiliar with the M200AU, we recommend  
that you read the overview Section 4 before proceeding. A diagram of the software menu  
trees is in Figure 5-1 and Figure 5-2.  
3.  
The M200AU requires about 30 minutes for all internal components to come up to  
temperature. During this time the ozone generator power is OFF until the membrane dryer  
has time to purge itself, therefore there will be no response from the instrument, even if span  
gas is coming in the sample port. During this time temperatures and other conditions are out  
of specification. Because many warning conditions could be displayed the software  
suppresses warning conditions for 30 minutes after power up. After 30 minutes, warning  
messages will be displayed until the respective warning conditions are within specifications.  
Use the CLR key on the front panel to clear warning messages.  
4.  
5.  
While waiting for instrument temperatures to come up, you can check for correct  
operation by using some of the M200AU's diagnostic and test features.  
Examine the TEST functions by comparing the values listed in Table 2-1 to those in the  
display. Remember that as the instrument warms up the values may not have reached their  
final values yet. If you would like to know more about the meaning and utility of each TEST  
function, refer to Table 9-1. Also, now is a good time to verify that the instrument was  
shipped with the options you ordered. Table 2-1 also contains the list of options. Section 6  
covers setting up the options.  
6.  
When the instrument is warmed up, re-check the TEST functions against Table 2-1. All  
of the readings should compare closely with those in the Table. If they do not, see Section  
9.1.1. The next task is to calibrate the analyzer. There are several ways to do a calibration,  
they are summarized in Table 7-1. For instruments not equipped with the external converter  
option, we recommend calibration with zero air and span gas coming in through the sample  
port. The procedure is:  
2-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Step 1 - Enter the expected NOx and NO span gas concentrations:  
Step Number Action  
Comment  
1.  
Press  
CAL-CONC-NOX  
This key sequence causes the M200AU to prompt for the  
expected NOx concentration. Enter the NOx span  
concentration value by pressing the key under each digit until  
the expected value is set.  
2.  
3.  
Press ENTR  
ENTR stores the expected NOx span value. This value will be  
used in the internal formulas to compute subsequent NOx  
concentration values.  
Press  
CAL-CONC-NO  
In the same CAL-CONC sub menu press the NO button and  
enter the expected NO span value, then ENTR. As before this  
value will be used in the internal formulas to compute the  
subsequent NO concentration values.  
4.  
5.  
Press EXIT-EXIT  
Returns instrument to SAMPLE mode.  
Press  
SETUP-RNGE-  
If necessary, you may want to change ranges. Normally the  
instrument is shipped in single range mode set at 500 ppb. We  
MODE-SING-ENTR recommend doing the initial checkout on the 500 ppb range.  
Press After SETUP-RNGE-SET, enter 500 and press ENTR. The  
SETUP-RNGE-SET instrument will now be in the 500 ppb range.  
6.  
7.  
Press EXIT  
Returns instrument to SAMPLE mode.  
2-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Step 2 - Calibrate the instrument:  
Zero/Span Calibration Procedure  
Step Number  
Action  
Comment  
1.  
Input Zero gas  
Allow Zero gas to enter the sample port on the rear of the  
instrument.  
2.  
Press CAL  
The M200AU enters the calibrate mode from sample mode.  
When the CAL button is pressed, the adaptive filter is  
activated. This allows the instrument to respond rapidly to  
concentration changes regardless of their magnitude.  
3.  
Wait 10 min  
Wait for reading to stabilize at the zero value. If you wait less  
than 10 minutes the final zero value may drift.  
4.  
5.  
Press ZERO  
Press ENTR  
The ZERO button will be displayed.  
Pressing ENTR actually changes the calculation equations and  
zeroes the instrument.  
6.  
7.  
Input Span Gas  
Wait 10 min  
Switch gas streams to span gas.  
Wait for reading to stabilize at the span value. If you wait less  
than 10 minutes the final span value may drift.  
8.  
Press SPAN  
The SPAN button should be displayed. If it is not, check the  
Troubleshooting Section 9.2.8 for instructions on how to  
proceed. In certain circumstances at low span gas  
concentrations (<100ppb), both the ZERO and SPAN buttons  
will appear.  
9.  
Press ENTR  
Press EXIT  
Pressing ENTR actually changes the calculation equations so  
that the concentration displayed is the same as the expected  
span concentration you entered above, thus spanning the  
instrument.  
10.  
Pressing EXIT returns the instrument to SAMPLE mode.  
2-8  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Step 3 - Review the quality of the calibration:  
Calibration Quality Check Procedure  
Step Number  
Action  
Comment  
1.  
Scroll the TEST  
function menu until  
the NOx SLOPE is  
displayed.  
The SLOPE value for NOx should be 1.0 ± 0.3. If the value is  
not in this range, check Section 7.8 or 9. If the SLOPE value  
is in the acceptable range, the instrument will perform  
optimally.  
2.  
Scroll the TEST  
function menu until  
the NO SLOPE is  
displayed.  
The SLOPE value for NO should be 1.0 ± 0.3. If the value is  
not in this range, check Section 7.8 or 9. If the SLOPE is in  
the acceptable range, the instrument will perform optimally.  
NOTE:  
The NO and NOx slopes should be equal within ± 0.1.  
4.  
5.  
Scroll the TEST  
function menu until  
The M200AU will display the OFFSET parameter for the NOx  
equation. This number should be near zero. A value of 0.0 mV  
the NOx OFFSET is –10 +150 mV indicates calibration in the optimal range. If the  
displayed.  
OFFSET value is outside this range, check Section 7 or 9 for  
procedures to correct the OFFSET value to near zero.  
Scroll the TEST  
function menu until  
the NO OFFSET is  
displayed.  
The instrument will now display the NO OFFSET value. It  
should also have a value near zero (0.0 mV –10 +150 mV).  
Step 4 - The M200AU is now ready to measure sample gas.  
2-9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Figure 2-4: Front Panel  
2-10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Figure 2-5: Assembly Layout  
2-11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 2-1: Final Test and Calibration Values  
Observed  
Value  
Test Values  
Units  
Nominal Range  
Reference Section  
RANGE  
ppb  
ppb  
5-2000  
5.3.4  
STABILITY  
0.001-2000  
9.1.1, 9.2.5,  
Table 9-1  
SAMP FLW  
OZONE FL  
PMT  
cc/min  
cc/min  
mV  
1000 ± 100  
80 ± 15  
9.3.7, Table 9-1  
9.3.6, 9.3.7  
9.3.8.1, Table 9-1  
Table 9-1  
9.3.8.5  
0-5000  
PREREACT  
HVPS  
mV  
0-1000  
V
450 - 900 constant  
2500 ± 200  
40 ± 1  
DCPS  
mV  
oC  
oC  
oC  
9.3.5, 9.3.4  
9.3.8.2  
RCELL TEMP  
BOX TEMP  
PMT TEMP  
MOLY TEMP  
RCEL PRES  
SAMP PRES  
8-48  
9.3.4.1  
-5 ± 1  
9.3.8.4  
oC  
315 ± 5  
9.3.4.1  
IN-Hg-A  
IN-Hg-A  
1 - 4 constant  
25 - 30 constant  
9.3.7  
9.3.7  
Electric Test & Optic Test  
Electric Test  
PMT Volts  
NO Conc  
NOx Conc  
mV  
2000 ± 500  
9.1.3.2  
9.1.3.2  
9.1.3.2  
PPB  
PPB  
1000 ± 250  
1000 ± 250  
Optic Test  
PMT Volts  
NO Conc  
mV  
2000 ± 1000  
1000 ± 500  
1000 ± 500  
9.1.3.3  
9.1.3.3  
9.1.3.3  
PPB  
PPB  
NOx Conc  
(table continued)  
2-12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 2-1: Final Test and Calibration Values (Continued)  
Observed  
Value  
Parameter  
Units  
Nominal Range  
Reference Section  
NO Span Conc  
NOx Span Conc  
NO Slope  
PPB  
PPB  
-
1 - 2000  
Table 7-3  
Table 7-3  
Table 7-18  
Table 7-18  
Table 7-18  
Table 7-18  
7.6.6, 5.2.2.6  
Table 9-1  
Table 9-1  
1 - 2000  
1.0 ± 0.3  
NOx Slope  
-
1.0 ± 0.3  
NO Offset  
mV  
mV  
%
-10 to +150  
-10 to +150  
0.96 - 1.02  
0.001 - 2000  
< 1 @ 400 ppb  
NOx Offset  
Moly Efficiency  
Stability at Zero  
Stability at Span  
PPB  
PPB  
Measured Flows  
Sample Flow  
Ozone Flow  
cc/min  
cc/min  
1000 ± 100  
80 ± 15  
9.3.7, Figure 9-8  
9.3.7, Figure 9-8  
Factory Installed Options  
Power Voltage/Frequency  
Option Installed  
Rack Mount, w/ Slides  
Rack Mount, w/ Ears Only  
Rack Mount, External Pump w/ Slides  
Rack Mount, External Pump w/o Slides  
Stainless Zero/Span Valves  
Current Loop - NOx Chan  
Current Loop - NO Chan  
Current Loop - NO2 Chan  
Current Loop - TST Chan  
4-20 mA  
0-20 mA  
Isolated  
Isolated  
Isolated  
Isolated  
Non-Isolated  
Non-Isolated  
Non-Isolated  
Non-Isolated  
4-20 mA  
4-20 mA  
4-20 mA  
0-20 mA  
0-20 mA  
0-20 mA  
PROM #  
Date  
Serial #  
Technician  
2-13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
INTENTIONALLY BLANK  
2-14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
3 SPECIFICATIONS, AGENCY APPROVALS,  
WARRANTY  
3.1 Specifications  
Ranges  
In 1ppb increments from 5ppb to 2000ppb  
independent ranges or autoranging  
<25 ppt RMS  
<0.25% RMS of reading above 50 ppb  
50 ppt RMS  
Noise at Zero1  
Noise at Span1  
Lower Detectable Limit1  
Zero Drift2  
<0.1 ppb / 24 hours  
<0.2 ppb / 7 days  
Span Drift2  
<0.5% FS or 50 ppt RMS whichever is greater / 7 days  
20 sec  
Lag Time  
Rise Time3  
95% in <50 sec6  
Fall Time3  
95% in <50 sec6  
Sample Flow Rate  
Linearity  
Precision  
1000 cc/min. ± 10%  
1% of full scale or ± 0.1 ppb whichever is greater  
0.5% of reading above 50 ppb  
Propylene rejection ratio > 20,000:1  
Ethylene rejection ratio > 40,000:1  
20-30oC within drift and noise specifications  
5-35oC safe operating range  
< 0.1% per oC  
0-95% RH non-condensing  
< 0.1% per V  
7" H x 17" W x 23.6" D (18cm x 43cm x 61cm)  
43 lbs (20 kg)  
Hydrocarbon Interference  
Temperature Range  
Temp Coefficient  
Humidity  
Voltage Coefficient  
Dimensions HxWxD  
Weight, Analyzer  
Weight, Ext Pump Pack  
Power, Analyzer  
Power, Analyzer4  
Power, Ext Pump  
Power, Ext Pump4  
Environmental4  
21 lbs (9.5 kg)  
100V~ 50/60 Hz, 120V~ 60 Hz, 230V~ 50 Hz, 2.5A, 125 watts  
230V~ 50 Hz, 2.5A  
110V~ 60 Hz, 220V~ 50 Hz, 240V~ 50Hz, 150 watts  
230V~ 50 Hz, 2.2A  
Installation Category (Over-voltage Category) II  
Pollution Degree 2  
Recorder Output  
Analog Resolution  
Status Option  
100 mV, 1, 5, 10V, isolated or non-isolated current loop  
1 part in 1024 of selected voltage or current range  
12 Status Outputs from opto-isolator  
ppb, ug/m3  
Measurement Units  
1.  
2.  
3.  
4.  
As defined by USEPA.  
At constant temperature and voltage.  
With adaptive filter, > 20 ppb change.  
Electrical rating for CE Mark compliance.  
3-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
3.2 EPA Equivalency Designation  
Teledyne Instruments Advanced Pollution Instrumentation Division, Model 200AU High  
Sensitivity Nitrogen Oxides Analyzer is designated as Reference Method Number RFNA-1194-  
099 as defined in 40 CFR Part 53, when operated under the following conditions:  
1.  
2.  
3.  
4.  
5.  
6.  
7.  
Range: Any range from 50 parts per billion (ppb) to 1 ppm.  
Ambient temperature range of 20 to 30oC.  
Line voltage range of 105-125 VAC, 60Hz; 220-240 VAC, 50Hz.  
With 5-micron TFE filter element installed in the internal filter assembly.  
Sample flow of 1000 ± 100 cc/min.  
Vacuum pump capable of 6"Hg Abs pressure @ 2 slpm or better.  
Software settings:  
A. Dynamic span  
B. Dynamic zero  
C. Cal-on-NO2  
D. Dilution factor  
E. AutoCal  
OFF  
OFF  
OFF  
OFF  
ON or OFF  
ON or OFF  
ON or OFF  
F. Independent range  
G. Autorange  
H. Temp/Pres compensation ON  
I. Converter Efficency 0.96 to 1.02  
Under the designation, the Analyzer may be operated with or without the following options:  
1.  
2.  
3.  
4.  
5.  
6.  
7.  
8.  
Rack mount with slides.  
Rack mount without slides, ears only.  
Rack mount for external pump w/o tray.  
Stainless steel zero/span valves.  
4-20mA, isolated outputs.  
Status outputs.  
RS-232 output.  
1 micron sample filter.  
3-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
3.3 Warranty  
WARRANTY POLICY  
Prior to shipment, Teledyne API equipment is thoroughly inspected and tested. Should  
equipment failure occur, Teledyne API assures its customers that prompt service and support  
will be available.  
COVERAGE  
After the warranty period and throughout the equipment lifetime, Teledyne API stands ready to  
provide on-site or in-plant service at reasonable rates similar to those of other manufacturers in  
the industry. All maintenance and the first level of field troubleshooting is to be performed by  
the customer.  
NON-TELEDYNE API MANUFACTURED EQUIPMENT  
Equipment provided but not manufactured by Teledyne API is warranted and will be repaired to  
the extent and according to the current terms and conditions of the respective equipment  
manufacturers warranty.  
GENERAL  
TELEDYNE API warrants each Product manufactured by Teledyne API to be free from defects  
in material and workmanship under normal use and service for a period of one year from the date  
of delivery. All replacement parts and repairs are warranted for 90 days after the purchase.  
If a Product fails to conform to its specifications within the warranty period, Teledyne API shall  
correct such defect by, in Teledyne API's discretion, repairing or replacing such defective  
Product or refunding the purchase price of such Product.  
The warranties set forth in this section shall be of no force or effect with respect to any Product:  
(i) that has been altered or subjected to misuse, negligence or accident, or (ii) that has been used  
in any manner other than in accordance with the instruction provided by Teledyne API or (iii)  
not properly maintained.  
THE WARRANTIES SET FORTH IN THIS SECTION AND THE REMEDIES  
THEREFORE ARE EXCLUSIVE AND IN LIEU OF ANY IMPLIED WARRANTIES OF  
MERCHANTABILITY, FITNESS FOR PARTICULAR PURPOSE OR OTHER  
WARRANTY OF QUALITY, WHETHER EXPRESSED OR IMPLIED. THE  
REMEDIES SET FORTH IN THIS SECTION ARE THE EXCLUSIVE REMEDIES FOR  
BREACH OF ANY WARRANTY CONTAINED HEREIN. TELEDYNE API SHALL  
NOT BE LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL DAMAGES  
ARISING OUT OF OR RELATED TO THIS AGREEMENT OF TELEDYNE API'S  
PERFORMANCE HEREUNDER, WHETHER FOR BREACH OF WARRANTY OR  
OTHERWISE.  
3-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
TERMS AND CONDITIONS  
All units or components returned to Teledyne API should be properly packed for handling and  
returned freight prepaid to the nearest designated Service Center. After the repair, the equipment  
will be returned, freight prepaid.  
3-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
4 THE M200AU NOX ANALYZER  
4.1 Principle of Operation  
The Teledyne API Model 200AU Analyzer is designed to measure the concentration of nitric  
oxide [NO], total oxides of nitrogen [NOx] and, by calculation, nitrogen dioxide [NO2].  
The instrument measures the light intensity of the chemiluminescent gas phase reaction of nitric  
oxide [NO] and ozone [O3] as follows:  
NO + O3 NO2* +O2  
NO2* NO2 + hv  
The reaction of NO with ozone results in electronically excited NO2 molecules as shown in the  
first equation above. The excited NO2 molecules release their excess energy by emitting a  
photon and dropping to a lower energy level as shown in the second equation. It has been shown  
that the light intensity produced is directly proportional to the [NO] concentration present.  
The Analyzer samples the gas stream and measures the [NO] concentration by digitizing the  
signal from the Analyzer's photomultiplier tube (PMT). A valve then routes the sample stream  
through a converter containing heated molybdenum to reduce any NOx present to NO by the  
following reaction:  
315o C  
3NOx + Mo 3NO + MoO3  
The Analyzer now measures the total NOx concentration. The [NOx] and [NO] values are  
subtracted from each other by the built-in computer, yielding the [NO2] concentration. In the  
third measurement phase, the instrument measures sample gas which has been mixed with ozone  
outside of the reaction cell. This Pre-Reactor allows the measurement of any hydrocarbon  
interferents present in the sample gas stream. The three results [NO], [NOx], and [NO2] are then  
further processed and stored by the computer yielding several instantaneous and long term  
averages of all three components.  
The software uses an adaptive filter to accommodate rapid changes in concentration. The  
algorithm monitors the rate of change in concentration for both the NO and NOx channels. When  
a change in concentration is detected, the software changes the sample filters to rapidly respond  
to the change. The filters are adjusted to minimize the errors introduced by the time delay  
between the NOx and NO channel measurements; this assures accurate NO2 measurements.  
When the rate of change decreases, the filters are lengthened to provide better signal/noise ratio.  
The parameters used to operate the adaptive filter have been tuned to match the electrical and  
pneumatic characteristics of the M200AU.  
4-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Figure 4-1: Block Diagram  
4-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
4.2 Operation Summary  
4.2.1 Sensor Module, Reaction Cell, Detector  
The sensor module (Figure 9-9) is where light from the chemilumenescent reaction is generated  
and detected. It is the most complicated and critical sub-assembly in the entire analyzer. It  
consists of the following assemblies and functions:  
1.  
2.  
3.  
4.  
5.  
The reaction cell and flow control module  
Reaction cell heater/thermistor  
PMT and high voltage power supply  
PMT cooler/cold block/heatsink/fan  
Preamp assembly:  
A. Preamp range control hardware  
B. HVPS control  
C. PMT cooler temp control  
D. Electric test electronics  
E. Optic test electronics  
4.2.2 Pneumatic Sensor Board  
The sensor board consists of 2 pressure sensors and a flow sensor. One pressure sensor measures  
the pressure in the reaction cell. The reaction cell is maintained at about 0.1 atmospheric  
pressure. The second pressure sensor measures the pressure just upstream of the reaction cell,  
which is near ambient pressure. From these two pressures the sample flow rate can be computed  
and is displayed as sample flow in the TEST menu. Finally, a solid state flow meter measures the  
ozone flow using a resistive bridge. Likewise, it is displayed as a TEST function.  
The M200AU displays all pressures in inches of mercury-absolute (in-Hg-A). Absolute pressure  
is the reading referenced to a vacuum or zero absolute pressure. This method was chosen so that  
ambiguities of pressure relative to ambient pressure can be avoided.  
4-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
NOTE  
On vacuum vs absolute pressure:  
Many vacuum gauges read relative to ambient pressure, therefore  
a reading of 25" of mercury (Hg) at sea level (which would give an  
absolute pressure of about 5" Hg in the reaction cell) would read  
only 20" Hg at high altitude sites. Therefore in this manual the vacuum  
specification of 5" Hg pressure is given as an absolute pressure  
- 5"Hg-A - reference against zero absolute pressure (a perfect vacuum)  
thus removing ambiguities for high altitude sites.  
4.2.3 Computer Hardware and Software  
CPU Board  
The M200AU Analyzer is operated by an V40 series micro computer. The computer's  
multitasking operating system allows it to do instrument control, monitor test points, provide  
analog output and provide a user interface via the display, keyboard and RS-232 port. These  
operations appear to be happening simultaneously but are actually done sequentially based on a  
priority queuing system maintained by the operating system. The jobs are queued for execution  
only when needed, therefore the system is very efficient with computer resources.  
The M200AU is a true computer based instrument. The microprocessor does most of the  
instrument control functions such as temperature control, valve switching. Data collection and  
processing are done entirely in the CPU with the final concentration values being sent to a D/A  
converter to produce the instrument analog output.  
The computer memory is divided into 3 sections: ROM memory contains the multi-tasking  
operating system code plus the instructions that run the instrument. The RAM memory is used to  
hold temporary variables, current concentration data and data acquisition system data. The  
EEPROM memory contains the instrument set-up variables such as range and instrument ID  
number. The EEPROM data is non-volatile so the instrument can lose power and the current set-  
up information is preserved.  
4.2.4 V/F Board  
The V/F board is multifunctional, consisting of A/D input channels, digital I/O channels, and  
analog output channels. Communication with the computer is via a STD bus interface. The  
computer receives all of the instrument data and provides all control functions through the V/F  
board.  
4-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
4.2.5 Front Panel  
The front panel of the M200AU is shown in Figure 2-4. The front panel consists of a 2 line  
display and keyboard, 3 status LED's and power switch. Communication with the display,  
keyboard, and status LED's is done via the computer's on-board parallel port. The M200AU was  
designed as a computer controlled instrument, therefore all major operations can be controlled  
from the front panel display and keyboard.  
The display consists of 2 lines of 40 characters each. The top line is divided into 3 fields, and  
displays information. The first field is the mode field. A list of operating modes is given in  
Table 5-8.  
The center field displays TEST values. The TEST functions allow you to quickly access many  
important internal operating parameters of the M200AU. This provides a quick check on the  
internal health of the instrument. The right hand field shows current concentration values of NO,  
NOx, and NO2. The display scrolls between the 3 values every 4 seconds.  
4.2.5.1 Keyboard  
The second line of the display contains eight fields. Each field defines the key immediately  
below it. By redefining the keys dynamically it is possible to simplify the instrument electronics  
and user interface.  
When entering data in the keyboard, if the entered value is not accepted, the M200AU will  
"beep" to notify the user that the value keyed in was not accepted. The original value remains  
unchanged.  
4.2.5.2 Status LED's  
At the right of the display there are 3 status LED's. They can be in three states, OFF, ON, and  
Blinking. The meanings of the LED's are given in Table 4-1.  
4-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 4-1: Front Panel Status LED's  
LED  
State  
Meaning  
Green  
On  
Off  
Monitoring normally, taking DAS data  
NOT monitoring, DAS disabled  
Blinking  
Monitoring, DAS in HOLDOFF mode (1)  
Yellow  
Red  
Off  
On  
Blinking  
Auto cal. disabled  
Auto/Dynamic cal. enabled  
Calibrating  
Off  
Blinking  
No warnings exist  
Warnings exist  
(1) This occurs during Calibration, DAS holdoff, power-up holdoff, and when in Diagnostic mode.  
4.2.5.3 Power Switch  
The power switch has two functions. The rocker switch controls overall power to the instrument,  
in addition it includes a circuit breaker. If attempts to power up the M200AU result in a circuit  
breaker trip, the switch automatically returns to the off position, and the instrument will not  
power up.  
4-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
4.2.6 Power Supply Module  
The Power supply module supplies AC and DC power to the rest of the instrument. It consists of  
a 4 output linear DC power supply and a 15 volt switching supply. In addition, it contains the  
switching circuitry to drive the DC operated valves and several switched AC loads to operate the  
Rx cell heater, the Moly heater, and the ozone generator.  
4.2.7 Pump, Valves, Pneumatic System  
A standard M200AU comes with 2 valves. The NO/NOx valve switches sample from either the  
sample inlet port or from the moly converter into the reaction cell. Each NO/NOx cycle, the Pre-  
reactor valve re-routes sample gas into the Pre-reactor volume which allows time for the NO-  
Ozone reaction to complete. The pneumatic timing of the Pre-reactor is set so that any  
hydrocarbon interferences are measured and eliminated from the signal.  
The M200AU is equipped with a high performance pump, capable of producing a reaction cell  
pressure of less than 4” Hg-A. See Figure 2-3 for hook-up information. A catalytic ozone  
scrubber protects the pump from the corrosive effects of ozone. See Section 4.2.8.  
4.2.8 Ozone Generator  
Because of the instability of ozone, it is necessary to generate this gas inside the analyzer. The  
ozone generation module consists of a high frequency switching AC supply and pulse  
transformer connected to a silent discharge tube. Air is supplied to the generator from a  
permeation type air drier. A complete description of its function and service requirements can be  
found in Section 9.3.6.  
Although there are dangerous high voltages generated in the ozone generator, they are isolated  
from the user by sealing the system in a single potted assembly. The dry air supply for the ozone  
generator uses a membrane drier to supply air with a dew point of 0o C or less. The exhaust side  
of the membrane is connected to the vacuum manifold at the rear of the instrument.  
Normal room air contains enough water vapor to damage the generator and components  
downstream. Because of this, the ozone generator may not turn on immediately at power up.  
A heated catalytic ozone scrubber, located in the molybdenum converter assembly removes  
excess ozone from the instrument exhaust. The delay is built into the instrument to allow the  
dryer to start operating and purge the system with dry air. Table 4-2 details the conditions for  
turning on the ozone generator.  
4-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 4-2: Ozone Generator Start-up Timing  
Time Since Last  
Ozone Gen State  
Program Action  
Power-up  
< 1 hour  
> 1 hour  
ON at power-up  
OFF at power-up  
Gen ON immediately after power-up.  
Wait 30 min, then turn gen ON.  
4.2.9 Molybdenum Converter – Ozone Scrubber  
The molybdenum converter is a stainless steel cartridge containing molybdenum chips heated to  
315° C. The converter's function is to reduce NOx to nitric oxide NO. The temperature control  
for this module is done by the Switch Card in the Power Supply Module using commands  
generated by the CPU. The temperature is measured by a type J thermocouple and is conditioned  
on the Status/Temp board. The analog voltage representing the Moly temperature is read by the  
V/F board and sent to the CPU where signals are generated for temperature control. The  
digitized voltage is translated to degrees for the TEST function on the front panel and for  
warnings.  
The molybdenum converter assembly also contains a catalytic ozone scrubber that removes  
excess ozone from the instrument exhaust.  
4-8  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
5 SOFTWARE FEATURES  
5.1 Index To Front Panel Menus  
The M200AU has 2 main operating modes, namely the SAMPLE and SETUP modes. The  
instrument is operating in the SAMPLE mode when it is measuring gas. The SETUP mode is  
used to create or change operating parameters such as range. Also in SETUP mode the  
instrument has extensive fault diagnosis tools. A list of M200AU operating modes is given in  
Table 5-8.  
The next several pages contain two different styles of indexes that will allow you to navigate the  
M200AU software menus. The first two pages show a "tree" menu structure to let you see at a  
glance where each software feature is located in the menu. The second menu contains a brief  
description of each key mnemonic and a reference to the section of the manual that describes its  
purpose and function in detail.  
5-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Figure 5-1: Sample Menu Tree  
5-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Figure 5-2: Setup Menu Tree  
5-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
5.1.1 Sample Menu  
Table 5-1: Sample Menu  
Menu Level  
Reference  
Section  
Level 1  
Level 2  
Level 3  
Level 4  
Description  
TEST  
TST>  
Test functions  
5.2.1, Table 9-1  
CAL  
Zero/Span calibration w/ gas  
through sample port  
5.2.2.1, 7.1  
CALZ  
CALS  
Zero calibration w/ zero gas  
from zero valve option  
5.2.2.2, 7.2, 7.3  
5.2.2.3, 7.2, 7.3  
5.2.2.2, 7.2, 7.3  
5.2.2.3, 7.2, 7.3  
7.1  
Span calibration w/ span gas  
from span valve option  
ZERO  
SPAN  
CONC  
Press ZERO then ENTR will  
zero analyzer  
Press SPAN then ENTR will  
span analyzer  
Expected NO/NOx span  
concentrations and Moly conv.  
efficiency setup  
NOX  
CONC  
Enter expected NOx span  
concentration  
5.2.2, Table 7-3  
5.2.2, Table 7-3  
NO  
CONC  
Enter expected NO span  
concentration  
5-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 5-2: Sample Menu  
Menu Level  
Reference  
Section  
Level 1  
Level 2  
Level 3  
Level 4  
Description  
CONV  
Sub-menu for converter  
efficiency setup and  
verification  
5.2.2.6, 7.6.6  
NO2  
CAL  
SET  
Expected NO2 concentration for  
converter efficiency calculation  
5.2.2.6, 7.6.6  
5.2.2.6, 7.6.6  
5.2.2.6, 7.6.6  
Automatic converter efficiency  
calibration and entry  
Set the converter efficiency  
manually  
MSG  
Displays warning messages  
Clears warning messages  
9.1.2  
CLR  
9.1.2  
SETUP  
The SETUP Menu - See next  
table  
Table 5-2  
5-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
5.1.2 Set-Up Menu  
Table 5-3: Setup Menu  
Setup Menu #1  
Reference  
Section  
Level 1  
Level 2  
Level 3  
Level 4  
Description  
CFG  
CFG is primarily used for  
showing special configuration  
options and factory special  
software  
5.3.1  
PREV,  
NEXT, LIST  
PREV, NEXT can be used to  
scroll through the  
5.3.1  
configuration list. LIST  
automatically scrolls the list  
AUTOCAL  
Automatic span check or  
calibration  
5.3.2, 6.4  
SEQUENCE  
Selects Sequence  
5.3.2, 6.4  
5.3.2, 6.4  
PREV-  
NEXT  
Scrolls display to select  
calibration sequence 1, 2, or 3  
MODE  
SET  
Selects mode of calibration  
(zero, span, zero-span) plus  
disable  
5.3.2, 6.4  
5.3.2, 6.4  
5.3.3  
For a given Sequence and  
Mode, sets timing and  
calibration attributes  
DAS  
Data Acquisition System  
(DAS) - keeps 1 to 1500  
minute averages of data  
(table continued)  
5-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 5-3: Setup Menu (Continued)  
Setup Menu #1  
Reference  
Section  
Level 1  
Level 2  
Level 3  
Level 4  
Description  
VIEW  
Select which DAS data  
collector to view  
5.3.3  
PREV-  
NEXT  
Scroll through data collectors  
CONC, PNUMTC, CAL  
DAT, STABILITY  
5.3.3  
EDIT  
Allows editing of the several  
attributes of a data collection  
channel.  
5.3.3  
5.3.3  
UP  
Displays the DAS data buffer  
- Move UP 1 average in the  
DAS data buffer  
UP10  
Move UP 10 averages in the  
DAS data buffer  
5.3.3  
5.3.3  
5.3.3  
5.3.3  
DOWN  
DOWN10  
PRNT  
Move down 1 average in the  
DAS data buffer  
Move DOWN 10 averages in  
the DAS data buffer  
Prints the setup parameters of  
a data collector to the RS-232  
port  
5-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 5-4: Setup Menu  
Setup Menu #2  
Reference  
Section  
Level 1  
Level 2  
MODE  
SET  
Level 3  
Level 4  
Description  
RANGE  
Range control menu  
5.3.4  
5.3.4  
Range mode select - Single,  
Autorange, Independent,  
Remote  
Sets range if mode is Single  
range  
5.3.4.1  
5.3.4.3  
5.3.4.3  
5.3.4.3  
5.3.4.2  
5.3.4.2  
NO  
NOx  
NO2  
LO  
Sets NO concentration range if  
indep ranges enabled  
Sets NOx concentration range if  
indep ranges enabled  
Sets NO2 concentration range if  
indep ranges enabled  
Sets low range if Autorange  
enabled  
HI  
Sets high range if Autorange  
enabled  
UNITS  
DIL  
Unit selection menu  
5.3.4.5  
5.3.4.5  
5.3.4.4  
PPB, UGM  
Select units that instrument uses  
Enter dilution factor if  
connected to stack dilution  
probe  
PASS  
Password enable/disable menu  
5.3.5  
5.3.5  
ON-OFF  
TIME  
Enable/disable password  
checking  
CLOCK  
Adjusts time on the internal  
time of day clock  
5.3.6  
5.3.6  
DATE  
Adjusts date on the internal  
time of day clock  
MORE  
Continue menus on next level  
down  
COMM  
RS-232 communications  
control menu  
5.3.8, 5.5  
5.3.8, 5.5  
5.3.8, 5.5  
BAUD  
ID  
Sets the BAUD rate to 300 -  
19,200  
Sets the instrument ID -  
(included on all RS-232  
messages)  
5-8  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 5-5: Setup Menu  
Setup Menu #3  
Reference  
Section  
Level 1  
Level 2  
Level 3  
Level 4  
Description  
VARS  
Internal variables  
5.3.9, 9.2  
5.3.9, 9.2  
PREV,  
NEXT,  
JUMP,  
EDIT  
PREV, NEXT scroll up and  
down through the VARS menu.  
JUMP will go to variable  
number selected, EDIT will  
allow editing of the selected  
variable.  
DIAG  
Diagnostic menu  
5.3.7, 9.1.3  
5.3.7, 9.1.3  
PREV,  
NEXT  
PREV, NEXT scroll up and  
down through the DIAG menu.  
SIG I/O  
Examines, changes analog and  
digital internal signals  
9.1.3.1  
9.1.3.5  
9.1.3.6  
9.1.1  
ANALOG  
OUT  
Writes a stepped analog output  
voltage to the 4 analog outputs  
D/A CAL  
Calibrates V/F board and the  
analog outputs  
TEST  
CHNL  
Routes several internal signals  
to TCHAN analog output - used  
for diagnosis  
OPTIC  
TEST  
Activates Optic Test feature  
Activates Electric Test feature  
Turns OFF/ON ozone generator  
9.1.3.3  
9.1.3.2  
ELEC  
TEST  
O3 GEN  
RS-232  
9.1.3.4  
9.1.3.7  
Writes test data to RS-232 port  
- used for diagnosis  
5-9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
5.2 Sample Mode  
5.2.1 Test Functions  
NOTE  
In any of the following TEST functions, if a value of  
XXXX is displayed, that indicates an off scale and  
therefore meaningless reading.  
To use the TEST functions to diagnose instrument faults, refer to Troubleshooting Section 9.1.  
Range  
This is the range of the instrument. Electronically, there is only one physical range. The range of  
the instrument is just the software expanding various portions of the single physical range to fill  
the selected analog output voltage range. In standard configuration there is one range for all 3  
outputs.  
Independent range option allows different ranges for each output. When enabled, there will be  
three range values displayed, NO, NOx and NO2.  
Auto range mode allows a low range and high range. The M200AU will automatically switch to  
the other range dynamically as concentration values require. The TEST values will show the  
range the instrument is currently operating in, and will dynamically display the alternate range as  
the range changes occur.  
NOTE  
Each of the range modes Single range, Auto range, and  
Independent ranges are mutually exclusive.  
Stability  
The instrument noise is computed using the standard deviation of the last 10 minutes of data,  
with the value being computed at the end of each NO/NOx cycle. It is computed for the NOx  
channel only. The stability value only becomes meaningful if sampling a constant concentration  
for more than 10 minutes. The value should be compared to the value observed in the factory  
check-out.  
The Stability reading on the front panel TEST functions is different than the STABIL reading  
found in the DAS. Check the DAS in Section 5.3.3 for more information on the STABIL - DAS.  
5-10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Sample Flow  
The SAMPLE FLOW test function is computed from the pressure measured up-stream of the  
sample flow orifice. The pressure down-stream of the orifice is also checked to assure the  
assumptions of the equation are valid. This will register variations in flow caused by changes in  
atmospheric pressure, but will not detect a plugged sample flow orifice. The nominal value is  
1000 ± 100 cc/min.  
Ozone Flow  
The OZONE FLOW test function is directly measured by a solid state flow meter. Variations in  
this value indicate variations in ozone flow. The nominal value for ozone flow is 80 ± 15 cc/min.  
PMT Voltage  
The PMT VOLTAGE measures the PMT signal at the output of the preamp board. The  
waveform of the PMT voltage can be complex, and vary from near 0 mV when zero gas is in the  
reaction cell to 5000 mV when a high concentration of NOx is being measured. If the PMT  
reading is consistently 5000 mV, that indicates an off-scale reading. Typical readings bounce  
around, which is normal.  
Normalized PMT Voltage  
Like the PMT Voltage TEST function above, the NORMALIZED PMT VOLTAGE measures  
the PMT signal at the output of the preamp board. The difference is that several normalization  
functions are applied to this signal before it is displayed. The most important is the temperature  
and pressure compensation factors. If NORM PMT is used as suggested in the Factory  
Calibration Procedure (Section 9.1.6) the M200AU will be correctly calibrated.  
PREREACT Voltage  
This test measurement is the Pre-reactor voltage. It indicates the most recent reading from the  
Pre-reactor circuit. The units are mV and readings up to 1000 are considered normal.  
High Voltage Power Supply (HVPS)  
The HVPS reading is a measure of the scaled-up HVPS programming voltage. The voltage used  
to set the HVPS output is generated on the Preamp board. Its value is between 0 and 1 volt,  
corresponding to a voltage of 0 to 1000 volts out of the HVPS. The HVPS front panel TEST  
measurement should be greater than 450 volts and will typically be 500-800 V.  
5-11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
DC Power Supply (DCPS)  
The DCPS voltage is a composite of the 5 and ± 15 VDC voltages in the Power Supply Module.  
This is meant to be a quick indicator to show if the PSM is working correctly. The nominal value  
is 2500 mV ± 200 mV.  
Reaction Cell Temperature  
This is a measurement of the temperature of the reaction cell. It is controlled by the computer to  
40 ± 1° C. Temperatures outside this range will cause the M200AU output to drift.  
Box Temperature  
This TEST function measures the temperature inside the chassis of the M200AU. The  
temperature sensor is located on the Status/Temp Board. Typically it runs 2 to 10° C higher than  
the ambient temperature.  
PMT Temperature  
The temperature of the PMT is closely controlled by a dedicated proportional temperature  
controller. The nominal set-point is -5 ± 1° C. Readings outside this range will cause instrument  
drift due to gain changes in the PMT detector.  
Block Temperature  
The temperature of the sample flow and ozone flow control orifice blocks. The nominal set-point  
is 40 ± 1° C. Readings outside this range will cause instrument drift.  
Moly Temperature  
The moly temperature is controlled by the computer. The nominal set-point is 315 ± 5° C. The  
temperature sensor inside the moly is a type-J thermocouple. The thermocouple amplifier is  
located on the STATUS/TEMP board. If the thermocouple breaks, the reading will go to 500° C  
and remove power from the heater.  
Reaction Cell Pressure  
The pressure in the reaction cell is measured by a solid state pressure sensor, which measures  
absolute pressure. This pressure will vary depending on several things.  
1.  
2.  
3.  
The type of pump attached to the analyzer.  
Variations in local weather will cause a ± 0.3in-Hg change in pressure.  
The altitude of the analyzer will cause the cell pressure to change.  
Nominal values are 1 to 4 in-Hg-A. typical reading is about 3.5 in-Hg-A.  
5-12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Sample Pressure  
The pressure in the sample inlet line is measured by a solid state pressure sensor which measures  
absolute pressure. This pressure typically runs 0.5" or so below atmospheric pressure due to the  
pressure drop in the sample inlet lines and valves.  
NOx, NO Slope and Offset Values  
The coefficients of 2 (NOx and NO) straight line equations (y = mx + b) determine the  
calibration of the M200AU. The slope parameter(m) can be thought of as a gain term which  
determines the steepness of the calibration curve. The offset parameter (b) compensates for  
differences in the background signal of the NO and NOx channels.  
The values of these parameters can be used to determine the quality of the calibration. These 4  
parameters contain valuable information about the quality and validity of the calibration. For  
example, the NO and NOx slope values should not differ by more than .1 from each other. Larger  
values indicate a flow imbalance such as a leak or problems with the molybdenum converter.  
Refer to Section 7.8 Calibration Quality for details on how to use these values.  
Time  
This is an output of the M200AU's internal time of day clock.  
5.2.2 CAL, CALS, CALZ, Calibration Functions  
The calibration and zero/span checking of the M200AU analyzer is treated in detail in Section 7,  
Table 7-1 summarizes types of calibration. The basic function of each of these keys is described  
here. When any of the CAL buttons are pressed (or whenever the instrument enters the CAL  
mode), the adaptive filter is activated. This allows the instrument to respond rapidly to  
concentration changes regardless of their magnitude.  
5.2.2.1 CAL, CALS, CALZ  
These keys control the calibration functions of the analyzer. In the CAL mode the analyzer can  
be calibrated without switching on the zero valve or the span valve. If the analyzer has the  
Zero/Span valve option, there will be CALZ and CALS buttons also. These buttons operate the  
Zero/Span valves. The setup of these options is covered in Section 6.3, and operation is  
explained in Section 7.  
5-13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
5.2.2.2 Zero  
Pressing the ZERO key along with ENTR will cause the instrument to adjust the OFFSET value  
of the internal formula so that the instrument reads zero. The M200AU allows zero adjustment  
over a limited range of signal levels, therefore the signal does not have to be exactly zero for the  
instrument to do a zero cal. The instrument will not, however, allow a zero cal on any signal  
level, therefore it is not possible to zero the instrument with high concentrations of span gas in  
the reaction cell. If the ZERO key does not come on as expected, check Section 9.2.9.  
5.2.2.3 Span  
Pressing the SPAN key along with ENTR will cause the instrument to adjust the SLOPE value of  
the internal formula so the instrument displays the span value. The expected NOx and NO span  
concentrations must be entered before doing a SPAN calibration. See Table 7-3.  
Like the Zero calibration, the Span cal cannot be done with any concentration of span gas. If the  
signal level is outside certain limits the, SPAN key will not be illuminated. If you encounter this  
condition see Section 9.2.8. It is also possible at low levels of span concentration that BOTH the  
ZERO and SPAN keys might be on, thus allowing you to either zero or span the instrument. In  
this case, care must be taken to perform the correct operation or the analyzer can become mis-  
calibrated.  
5.2.2.4 NO, NOx Cal Concentration  
Before the M200AU can be spanned, it is necessary to enter the expected span concentrations for  
NO and NOx. This is done by using CAL-CONC-NOX or CAL-CONC-NO keys for NOx and  
NO span concentrations, respectively. Concentration values from 1 to 2000 ppb are valid  
5-14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
5.2.2.5 Formula Values  
The slope and offset terms should be checked after each calibration. The values for these terms  
contain valuable information about the internal health of the analyzer. The range of acceptable  
values and their meanings is given in Section 7.8.  
To compute the NOx and NO concentrations, the formula for a straight line is used.  
Where:  
y = the NOx or NO concentration  
m = the slope  
x = the conditioned PMT tube output  
b = the offset  
y = mx + b  
In comparison with analog analyzers the slope term is equivalent to the "span pot" and the b term  
is equivalent to the "zero pot". Again, like an analog analyzer there is only a limited range of  
adjustment allowed for either term, and there are consequences of having the values near the  
high or low limits of their respective ranges.  
The x term is the conditioned PMT signal. PMT signal is adjusted for the Pre-reactor  
background, range, temperature, and pressure.  
The offset (b) term is the total background light with the Pre-reactor term subtracted out.  
The Pre-reactor term measures detector dark current, amplifier noise, and ozone generator and  
hydrocarbon background.  
After every zero or span calibration, it is very important to check the QUALITY of the  
calibration. The calibration of the M200AU involves balancing several sections of electronics  
and software to achieve an optimum balance of accuracy, noise, linearity and dynamic range.  
See Section 7.8 for the calibration quality check procedure.  
5-15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
5.2.2.6 Automatic Converter Efficiency Compensation  
The M200AU can automatically compensate the NOx and NO2 readings for the molybdenum  
converter efficiency. There are 2 ways to enter the converter efficiency into the instrument. The  
first is to just type in the converter efficiency using the CAL-CONC-MOLY-SET menu. The  
second method is to have the M200AU compute the efficiency using the CAL-CONC-MOLY-  
CAL menu. See the Calibration Section 7.6.6.1 Molybdenum Converter Efficiency for details.  
To disable the compensation, press CAL-CONC-MOLY-SET and enter 1.0000 as the efficiency.  
Factory default is 1.0000.  
5.3 Set-Up Mode  
5.3.1 Configuration Information (CFG)  
This menu item will tell if the installed software has factory special features or other non-  
standard features. If you call Teledyne API service you may be asked for information from this  
menu.  
5.3.2 Automatic Calibration (AutoCal)  
The AutoCal feature allows the M200AU to automatically operate the Zero/Span Valve option to  
periodically check its calibration. Information on setting up AutoCal is in Section 6.3.  
5.3.3 Data Acquisition System (DAS)  
The M200AU contains a flexible and powerful built in data acquisition system (DAS) that  
enables the analyzer to store concentration data as well as diagnostic parameters in its battery  
backed memory. This information can be viewed from the front panel or printed out through the  
RS-232 port. The diagnostic data can be used for performing “Predictive Diagnostics” and  
trending to determine when maintenance and servicing will be required.  
The logged parameters are stored in what are called “Data Channels.” Each Data Channel can  
store multiple data parameters. The Data Channels can be programmed and customized from the  
front panel. A set of default Data Channels has been included in the M200AU software. For  
more information on programming custom Data Channels, a supplementary document containing  
this information can be requested from Teledyne API.  
5-16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
5.3.3.1 Data Channels  
The function of the Data Channels is to store, report, and view data from the analyzer. The data  
may consist of NO, NOx, NO2 concentration, or may be diagnostic data, such as the sample flow  
or reaction cell pressure.  
The M200AU comes pre-programmed with a set of useful Data Channels for logging  
concentration and predictive diagnostic data. The default Data Channels can be used as they are,  
or they can be changed by the user to fit a specific application. They can also be deleted to make  
room for custom user-programmed Data Channels.  
The data in the default Data Channels can be viewed through the SETUP-DAS-VIEW menu.  
Use the PREV and NEXT buttons to scroll through the Data Channels and press VIEW to view  
the data. The last record in the Data Channel is shown. Pressing PREV and NEXT will scroll  
through the records one at a time. Pressing NX10 and PV10 will move forward or backward 10  
records. For Data Channels that log more than one parameter, such as PNUMTC, buttons labeled  
<PRM and PRM> will appear. These buttons are used to scroll through the parameters located  
in each record.  
5-17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
The function of each of the default Data Channels is described below:  
Samples NOx, NO and NOy concentration data at one minute intervals and  
stores an average every hour with a time and date stamp. Readings during  
calibration and calibration hold off are not included in the data. The last 800  
hourly averages are stored.  
CONC:  
Collects sample flow and sample pressure data at five minute intervals and  
stores an average once a day with a time and date stamp. This data is useful for  
monitoring the condition of the pump and critical flow orifice(sample flow) and  
the sample filter(clogging indicated by a drop in sample pressure) over time to  
predict when maintenance will be required. The last 360 daily averages (about 1  
year) are stored.  
PNUMTC:  
Logs new slope and offset every time a zero or span calibration is performed,  
also records the sample concentration reading just prior to performing a  
calibration.  
CALDAT:  
STABIL:  
Logs the standard deviation of the last 50 minutes of NOx data, with readings  
taken 2 minutes apart(i.e. 25 readings taken 2 minutes apart.) The collector  
operates in “sliding window” fashion with the oldest reading being deleted and  
a new reading added every 2 minutes. A new value of STABIL is added to the  
data buffer every 2 minutes. A time and date stamp is recorded for every data  
point logged.  
NOTE:  
This Data Channel collects data based on an event (a calibration) rather than a  
timer. This Data Channel will store data from the last 200 calibrations. This  
does not represent any specific length of time since it is dependent on how often  
calibrations are performed. As with all Data Channels, a time and date stamp is  
recorded for every data point logged.  
This is the EPA definition of noise used for equivalency testing. The collector provides a  
continuous readout of the noise. In order for the data to be meaningful, the instrument must  
sample a constant concentration of gas for at least 50 minutes.  
5-18  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 5-6: DAS Data Channel Editing  
Step  
Action  
Comment  
1.  
2.  
3.  
4.  
Enter DAS menu to edit Data Channels  
Select Data Channel to edit  
Press SETUP-DAS-EDIT  
Press PREV/NEXT  
Press EDIT  
Enter the Edit menu for the selected Data Channel  
Scroll through Data Channel properties until RS-232  
REPORT: OFF is displayed  
Press SET> (5 times)  
5.  
6.  
7.  
8.  
Edit selected setup property  
Change RS-232 REPORT property  
Accepts change  
Press EDIT  
Toggle OFF to ON  
Press ENTR  
Exits back to sample menu  
Press EXIT (4 times)  
5.3.4 Range Menu  
The instrument operates on any full scale range from 5 to 2000 ppb. The range is the  
concentration value that equals the maximum voltage output on the rear panel of the instrument.  
The front panel will read the concentration anywhere from 0 to 2000 ppb regardless of the range  
selected.  
NOTE  
Only one of the following range choices can be active at any one time.  
There are 3 range choices:  
1.  
2.  
3.  
Single Range  
Auto Range  
Independent Ranges  
5.3.4.1 Single Range  
This range option selects a single range for all output channels (NO, NOx, NO2) of the M200AU.  
To select Single Range press SETUP-RNGE-MODE-SING, then press ENTR. To set the value  
for the range press SETUP-RNGE-SET, enter the full scale range desired from 5 ppb to 2000  
ppb, then press ENTR.  
5-19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
5.3.4.2 Auto Range  
Auto Range allows the NO, NOx, NO2 outputs to automatically range between a low range and a  
high range. There is only one low range and one high range for all outputs. The Hi range mode is  
signaled by a bit on the STATUS option, see Table 5-9. When the instrument output increases to  
98% of the low range value, it will Auto Range into Hi range. In Hi range, when the output  
decreases to 75% of low range, it will change to the lower range. If you select a Hi range that is  
less than Low range, the M200AU will remain locked in Low range and behave as a Single  
Range instrument.  
To set up Auto Range press SETUP-RNGE-MODE-AUTO, then press ENTR. To set the values  
press SETUP-RNGE-SET. The M200AU will prompt you for LO, then HI which are the lower  
and upper ranges of Auto Range. Key in the values desired, then press ENTR.  
5.3.4.3 Independent Ranges  
Independent Ranges allows you to select different ranges for NO, NOx, and NO2.  
To set up Independent Ranges press SETUP-RNGE-MODE-IND, then press ENTR. To set the  
values press SETUP-RNGE-SET. The M200AU will prompt you for the range of NO, NOx and  
NO2 channels. Key in the desired range for each channel, press ENTR after each value.  
5.3.4.4 Concentration Units  
The M200AU can display concentrations in ppb, ug/m3 units. Concentrations displayed in ug/m3  
use 0° C, 760 mm-Hg for STP. Consult your local regulations for the STP used by your agency.  
The following equations give approximate conversions:  
NO ppb x 1.34 = NO ug/m3  
NO ppm x 1.34 = NO mg/m3  
NO2 ppb x 2.05 = NO2 ug/m3  
NO2 ppm x 2.05 = NO2 mg/m3  
NH3 ppb x 0.76 = NH3 ug/m3  
NH3 ppm x 0.76 = NH3 mg/m3  
To change the current units press SETUP-RNGE-UNIT from the SAMPLE mode and select the  
desired units.  
5-20  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
NOTE  
You should now reset the expected span concentration values in the new  
units and re-calibrate the instrument using one of the methods in Section 7.  
Changing units affects all of the RS-232 values, all of the  
display values, and all of the calibration values.  
Example:  
If the current units are in ppb and the span value is 400 ppb, and  
the units are changed to ug/m3 the span value is NOT re-calculated  
to the equivalent value in ug/m3. Therefore the span value now  
becomes 400 ug/m3 instead of 400 ppb.  
5.3.4.5 Recorder Offset  
If necessary, the analog outputs can be biased. This can be used to offset the output voltage of  
each channel ±10% of the current output voltage setting. It is intended for recorders that cannot  
show slightly negative readings. It can also be used to bias the input to a datalogger to offset  
small external ground loop voltages that are sometimes present in monitoring systems  
The offset is set in the V/F calibration menu. Press SETUP-MORE-DIAG, scroll to A/D  
CALIBRATION, then press ENTR. Select CFG-SET-OFFSET, then enter the desired offset and  
press ENTR. Press EXIT to return to the SAMPLE mode.  
5.3.5 Password Enable  
If password protection is enabled, a password is required to access calibration or setup menus. In  
the VARS menu a password is always required. To enable passwords press SETUP-PASS-ON.  
A list of passwords is in Table 5-7.  
Table 5-7: Calibrate, Setup Passwords  
Password Usage  
Calibration Password  
Setup Password  
Password  
512, 101  
818, 101  
Use to get into CAL menus  
Use to get into SETUP menus  
5-21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
5.3.6 Time of Day Clock  
The instrument has an internal time of day clock. The time of day can be set by pressing SETUP-  
CLOCK-TIME and entering the time in 24hr format. In a similar manner the date can be entered  
by pressing SETUP-CLOCK-DATE and entering the date in a dd-mmm-yy format.  
If you are having trouble with the clock running slow or fast, the speed of the clock can be  
adjusted by selecting the CLOCK_ADJ variable in the SETUP-MORE-VARS menu. The units  
of CLOCK_ADJ are seconds per day.  
5.3.7 Diagnostic Mode  
The M200AU Diagnostic Mode allows additional tests and calibrations of the instrument. These  
features are separate from the TEST functions because each DIAG function has the ability to  
alter or disable the output of the instrument. While in DIAG mode no data is placed in the DAS  
averages. Details on the use of Diagnostic mode are in Section 9.1.3.  
5.3.8 Communications Menu  
The COMM menu allows the RS-232 BAUD rate to be set. To set the BAUD rate press SETUP-  
MORE-COMM-BAUD, select the appropriate BAUD rate, then press ENTR.  
The instrument ID number can also be set. This ID number is attached to every RS-232 message  
sent by the M200AU. To set the ID press SETUP-MORE-COMM-ID and enter a 4 digit number  
from 0000-9999, then press ENTR  
5.3.9 Variables Menu (VARS)  
This menu enables you to change the settings on certain internal variables. The VARS Table 9-5  
is located in the Troubleshooting Section 9.1.4.  
5.3.10 M200AU Operating Modes  
The M200AU has 2 main operating modes that were discussed earlier in this section, namely  
SAMPLE and SETUP modes. In addition, there are other modes of operation when the  
instrument is being diagnosed or calibrated. A list of M200AU operating modes is given in  
Table 5-5.  
5-22  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 5-8: M200AU Operating Modes  
Mode  
Description  
ZERO CAL D  
ZERO CAL A  
ZERO CAL R  
ZERO CAL M  
SPAN CAL D  
SPAN CAL A  
SPAN CAL R  
SPAN CAL M  
M-P CAL  
Automatic dynamic zero calibration  
Automatic zero calibration  
Remote zero calibration  
Manual zero calibration  
Automatic dynamic span calibration  
Automatic span calibration  
Remote span calibration  
Manual span calibration  
Manual multi-point calibration  
Electrical diagnostic test  
DIAG ELEC  
DIAG OPTIC  
DIAG OZONE  
DIAG AOUT  
DIAG  
Optical diagnostic test  
Ozone generator diagnostic test  
D/A output diagnostic test  
Main diagnostic menu  
DIAG I/O  
Signal I/O diagnostic  
DIAG RS232  
SETUP x.x  
RS232 output diagnostic  
Setup mode (x.x is software version)  
Sampling; automatic dynamic zero and span calibration enabled  
Sampling; automatic dynamic zero calibration enabled  
Sampling; automatic dynamic span calibration enabled  
Sampling; automatic cal. enabled  
Sampling; automatic cal. disabled  
SAMPLE ZS  
SAMPLE Z  
SAMPLE S  
SAMPLE A  
SAMPLE  
5.4 Status Output  
The status output is an option that reports Analyzer conditions via contact closures on the rear  
panel. The closures are available on a 50 pin connector on the rear panel. The contacts are NPN  
transistors which can pass 50 ma of direct current. The pin assignments are listed in Table 5-9.  
5-23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 5-9: Status Output Pin Assignments  
Output #  
Pin #  
1,2  
Definition  
Condition  
1
2
ZERO CAL  
SPAN CAL  
FLOW ALARM  
TEMP ALARM  
DIAG MODE  
POWER OK  
SPARE  
CLOSED IN ZERO CAL  
CLOSED IN SPAN CAL  
3,4  
3
5,6  
CLOSED IF FLOW WARNING  
CLOSED IF ANY TEMP WARNING  
CLOSED IN DIAG MODE  
CLOSED IF SYSTEM POWER OK  
4
7,8  
5
9,10  
11,12  
13,14  
15,16  
17,18  
19,20  
21,22  
23,24  
6
7
8
SPARE  
9
SPARE  
10  
11  
12  
AUTORANGE - HI  
SYSTEM OK  
RX CELL PRESS  
CLOSED IF IN HIGH RANGE  
CLOSED IF NO FAULTS PRESENT  
CLOSED IF ABS PRES > 15" HG  
The Status Board schematic can be found in the Appendix Drawing 01087.  
5.5 RS-232 Interface  
The RS-232 communications protocol allows the instrument to be connected to a wide variety of  
computer based equipment. The interface provides two basic functions in the M200AU.  
1.  
2.  
First is a comprehensive command interface for operating and diagnosing the analyzer.  
The interface can also provide an audit trail of analyzer events. In this function the port  
sends out messages about instrument events like calibration or warning messages. If these  
messages are captured on a printer or remote computer, they provide a continuous audit trail  
of the analyzers operation and status.  
5-24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
5.5.1 Setting Up the RS-232 Interface  
The baud rate is set from the front panel by SETUP-MORE-COMM-BAUD. Select the baud rate  
appropriate for your application, 300, 1200, 2400, 4800, 9600, or 19,200. It is important to note  
that the other device must have identical settings in order for the communications to work  
correctly.  
Second is physical wiring of the analyzer to the other unit. We have incorporated into the  
Analyzer LED's that signal the presence of data on the communications lines, and also switches  
to easily re-configure the analyzer from DCE to DTE if necessary. In addition the front panel  
diagnostics allow test data streams to be sent out of the port on command. This flexibility and  
diagnostic capability should simplify attaching our equipment to other computers or printers. If  
problems occur, see the Troubleshooting Section 9.3.2.  
Setup from the Front Panel  
There are 2 additional RS-232 setups that can be done via the front panel.  
1.  
2.  
Set the instrument ID number by SETUP-MORE-COMM-ID, and enter a 4 digit number  
from 0000-9999. This ID number is part of every message transmitted from the port.  
Set the RS-232 mode bit field in the VARS menu. To get to the variable press, SETUP-  
MORE-VARS, then ENTR and scroll to RS232_MODE, then press EDIT. The possible  
values are in Table 5-10.  
Table 5-10: RS-232 Port Setup - Front Panel  
Decimal Value  
Description  
1
2
Turns on quiet mode (messages suppressed)  
Places analyzer in computer mode (no echo of chars)  
Enables Security Features (Logon, Logoff)  
Enables API protocol and setup menus  
Enable alternate protocol  
4
8
16  
32  
Enable multidrop protocol  
NOTE  
To enter the correct value, ADD the decimal values of the features you  
want to enable. For example if LOGON and front panel RS-232  
menus are desired, the value entered would be 4 + 8 = 12.  
5-25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Security Feature  
The RS-232 port is often connected to a public telephone line which could compromise  
instrument security. If the LOGON feature is implemented, the port has the following attributes:  
1.  
2.  
3.  
A password is required before the port will operate.  
If the port is inactive for 1 hour, it will automatically LOGOFF.  
Repeat attempts at logging on with incorrect passwords will cause subsequent logins  
(even with the correct password) to be disabled for 1 hour.  
4.  
5.  
If not logged on, the only command that is active is the '?'.  
The following messages will be given at logon.  
LOG ON SUCCESSFUL  
LOG ON FAILED  
Correct password given  
Password not given or incorrect  
Logged off  
LOG OFF SUCCESSFUL  
The RS-232 LOGON feature must be enabled from the front panel by setting bit 4 of the  
RS232_MODE variable in the VARS menu, see Table 9-5. Once the feature is enabled, to logon  
type:  
LOGON 940331  
940331 is the default password. The password can be changed to any number from 0 to 999999  
by the variable RS232_PASS. To change the password enter the command  
V RS232_PASS=NNNNNN  
which sets the password to the value NNNNNN.  
Protocol of Port Communication  
The RS-232 interface has two protocols of communication, because if the port is attached to a  
computer it needs to have different characteristics than if used interactively. Consequently, there  
are two primary styles of operation: terminal mode and computer mode.  
When an operator is communicating with the analyzer via a terminal, the analyzer should be  
placed into TERMINAL MODE, which echoes keystrokes, allows editing of the command line  
using the backspace and escape keys, and allows recall of the previous command. When a host  
computer or data logger is connected to the analyzer, it should be placed into COMPUTER  
MODE, which does not echo characters received or allow the special editing keys. See  
Table 5-11 for relevant commands.  
5-26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 5-11: RS-232 Switching From Terminal Mode to Computer Mode  
Key  
Function  
Control-T (ASCII 20 decimal)  
Control-C (ASCII 3 decimal)  
Switch to terminal mode (echo, edit)  
Switch to computer mode (no echo, no edit)  
If the command line doesn't seem to respond to keystrokes or commands, one of the first things  
you should do is send a Control-T to switch the command line interface into terminal mode.  
Also, some communication programs remove CTRL-T and CTRL-C characters from the byte  
stream, therefore these characters will not be sent to the analyzer. Check your communications  
program owners manual.  
Entering Commands in Terminal Mode  
In terminal mode, all commands must be terminated by a carriage return; commands are not  
processed until a carriage return is entered. While entering a command you may use the editing  
keys shown in Table 5-12.  
Table 5-12: RS-232 Terminal Mode Editing Keys  
Key  
Function  
CR (carriage return)  
BS (backspace)  
ESC (escape)  
Execute command  
Backspace one character to the left  
Erase entire line  
Commands are not case-sensitive; you should separate all command elements (i.e. keywords,  
data values, etc.) by spaces.  
Words such as T, SET, LIST, etc. are called keywords and are shown on the help screen in  
uppercase, but they are not case-sensitive. You must type the entire keyword; abbreviations are  
not accepted.  
NOTE  
To open the help screen, Type "?" and press the Enter key.  
5-27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
5.5.2 Command Summary  
The information contained in the rest of this section covers commonly used commands that are  
required to operate the instrument from a remote terminal. If you are going to be writing  
computer programs to communicate with the M200AU (i.e. operating the port in COMPUTER  
MODE) we suggest that you order a supplementary manual "The RS-232 Interface", Teledyne  
API part number 01350. This manual describes additional features of the port.  
The Teledyne API RS-232 interface protocol has a multidrop capability. This is why an optional  
ID number is permitted for all commands. If you don’t include the ID number in the command,  
all of the instruments connected to the RS-232 interface will respond. If you include the ID  
number in the command, only the instrument whose ID number matches will execute the  
command.  
5-28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 5-13: RS-232 Command Summary  
Commands  
? [id]  
Definition  
Print help screen. ID is an optional instrument ID number.  
Print all active test messages  
T [id] LIST  
T [id] LIST name or  
T [id] name  
Print single test message "name" from Table 5-16  
W [id] LIST  
Print all active warnings  
W [id] CLEAR name or  
W [id] name  
Clear single warning message "name" from Table 5-17  
C [id] command  
D [id] LIST  
Execute calibration "command" from Table 5-18  
Prints all I/O signal values  
D [id] name  
Print single I/O signal value/state  
Sets I/O signal to new "value"  
Lists diagnostic test names  
D [id] name=value  
D [id] LIST NAMES  
D [id] ENTER name  
D [id] EXIT  
Enters and starts 'name' diagnostic test  
Exits diagnostic mode  
D [id] RESET  
Resets analyzer(same as power-on)  
D [id] RESET RAM  
System reset, plus erases RAM. Initializes DAS, NO, NOx, NO2 conc  
readings, calib not affected.  
D [id] RESET EEPROM  
System reset, plus erases EEPROM (RESET RAM actions + setup  
variables, calibration to default values). Restores all factory defaults.  
D [id] PRINT  
Prints properties for all data channels (DAS)  
D [id] PRINT "name”  
Prints properties for single data channel. Quotes around name are  
required.  
D [id] REPORT "name"  
[RECORDS=number]  
[COMPACT|VERBOSE]  
Prints DAS records for a data channel. Quotes around name are  
required. Parameters in brackets are optional.  
V [id] LIST  
Print all setup variable names and values  
Print individual setup variable value  
Sets setup variable to new "value"  
Print analyzer configuration  
V [id] name  
V [id] name=value  
V [id] CONFIG  
V [id] MODE  
Print current analyzer mode  
5-29  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 5-14: RS-232 Command Summary  
Terminal Mode Editing Keys  
Definition  
BS  
Backspace  
ESC  
Erase line  
CR  
Execute command  
Switch to computer mode  
Definition  
^C  
Computer Mode Editing Keys  
LF  
Execute command  
Switch to terminal mode  
Definition  
^T  
Security Features  
LOGON [id] password  
LOGOFF [id]  
Establish connection to analyzer  
Disconnect from analyzer  
General Output Message Format  
Reporting of status messages for use as an audit trail is one of the two principal uses for the RS-  
232 interface. You can effectively disable the asynchronous reporting feature by setting the  
interface to quiet mode. All messages output from the analyzer (including those output in  
response to a command line request) have the format:  
X DDD:HH:MM IIII MESSAGE  
X is a character indicating the message type, as shown in the table below.  
DDD:HH:MM is a time-stamp indicating the day-of-year (DDD) as a number from 1 to 366, the  
hour of the day (HH) as a number from 00 to 23, and the minute (MM) as a number from 00 to  
59.  
IIII is the 4-digit machine ID number.  
MESSAGE contains warning messages, test measurements, DAS reports, variable values, etc.  
The uniform nature of the output messages makes it easy for a host computer to parse them.  
5-30  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 5-15: RS-232 Interface Command Types  
First Character  
Message Type  
Calibration  
Diagnostic  
C
D
T
Test measurement  
Variable  
V
W
Warning  
There are 5 different types of messages output by the M200AU. They are grouped below by type  
in Table 5-13 to Table 5-21. The meanings of the various messages are discussed elsewhere in  
the manual. The TEST, DIAGNOSTIC and WARNING messages are discussed in Sections 9.1  
and 9.2. DAS and VARIABLES are discussed in Section 5.3.5 and 5.3.9. CALIBRATE is  
discussed in Section 7.  
5-31  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
5.5.3 TEST Commands and Messages  
Table 5-16: RS-232 Test Messages  
Name  
RANGE2  
Message  
Description  
RANGE=xxxxx PPB3  
NOX RNG=xxxxx PPB3  
NO RNG=xxxxx PPB3  
NO2 RNG=xxxxx PPB3  
NOX STB=xxxx.xx PPB  
SAMP FLW=xxx CC/M  
OZONE FL=xxxx CC/M  
PMT=xxxxxx MV  
Analyzer range  
NOXRANGE1  
NORANGE1  
NO2RANGE1  
STABILITY  
SAMPFLOW  
OZONEFLOW  
PMT  
Indep. Range for NOx channel  
Indep. Range for NO channel  
Indep. Range for NO2 channel  
Std. Deviation NOx conc values.  
Sample flow rate  
Ozone flow rate  
PMT output  
NORMPMT  
PRE-REACTOR  
HVPS  
NORM PMT=xxxxxx MV  
PREREACT=xxxxx MV  
HVPS=xxxxx V  
Normalized PMT output  
Pre-reactor filter value  
High voltage power supply  
DC power supply  
DCPS  
DCPS=xxxxxx MV  
RCELL TEMP=xxx C  
BOX TEMP=xxx C  
PMT TEMP=xxx C  
MOLY TEMP=xxx C  
RCEL=xxx.x IN-HG-A  
SAMP=xxx.x IN-HG-A  
NOX SLOPE=xxxxx  
NOX OFFS=xxxxx  
NO SLOPE=xxxxxx  
NO OFFS=xxxxxx  
RCELLTEMP  
BOXTEMP  
PMTTEMP  
CONVTEMP  
RCELLPRESS  
SAMPPRESS  
NOXSLOPE  
NOXOFFSET  
NOSLOPE  
NOOFFSET  
NO2CONC  
NOXCONC  
NOCONC  
Reaction cell temperature  
Internal box temperature  
PMT temperature  
Molycon temperature  
Rx cell pressure  
Sample pressure  
NOx slope parameter  
NOx offset parameter  
NO slope parameter  
NO offset parameter  
NO2=xxxxx PPB  
Instantaneous NO2 concentration  
Instantaneous NOx concentration  
Instantaneous NO concentration  
Test channel diagnostic output  
Time of day  
NOX=xxxxx PPB  
NO=xxxxx PPB  
TESTCHAN  
CLOCKTIME  
TEST=xxxxx MV  
TIME=HH:MM:SS  
1Displayed when independent range is enabled.  
2Displayed when single or auto range is enabled.  
3Depends on which units are currently selected.  
5-32  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
The T command lists TEST messages. Examples of the T command are:  
T LIST  
Lists all active test messages  
T LIST ALL  
T CONVTEMP  
T LIST NOX  
T NOX  
Lists all test messages  
Prints the temperature of the moly converter  
Prints NOx concentration message  
Prints NOx concentration message  
5.5.4 WARNING Commands and Messages  
Table 5-17: RS-232 Warning Messages  
Name  
Message  
Description  
WSYSRES  
SYSTEM RESET  
Analyzer was reset/powered on  
RAM was erased  
WRAMINIT  
RAM INITIALIZED  
WSAMPFLOW  
WOZONEFLOW  
WRCELLPRESS  
WBOXTEMP  
WRCELLTEMP  
WCONVTEMP  
WPMTTEMP  
WPREREACT  
SAMPLE FLOW WARN  
OZONE FLOW WARNING  
RCELL PRESS WARN  
BOX TEMP WARNING  
RCELL TEMP WARNING  
MOLY TEMP WARNING  
PMT TEMP WARNING  
PRACT WARN XXX.X MV  
Sample flow out of spec.  
Ozone flow out of spec.  
Rx cell pressure out of spec.  
Box temp. out of spec.  
Reaction cell temp. out of spec.  
Molycon temp. out of spec.  
Molycon temp. out of spec.  
Pre-reactor filter receive a reading out  
of limit spec.  
WHVPS  
HVPS WARNING  
High voltage out of spec.  
WDCPS  
DCPS WARNING  
DC Voltage out of spec.  
WOZONEGEN  
WDYNZERO  
WDYNSPAN  
WVFDET  
OZONE GEN OFF  
CANNOT DYN ZERO  
CANNOT DYN SPAN  
V/F NOT DETECTED  
Ozone Generator is off  
Dynamic zero cal. out of spec.  
Dynamic span cal. out of spec.  
V/F board not installed or broken  
5-33  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Whenever a warning message is reported on the analyzer display, if the RS-232 interface is in  
the normal mode(i.e. not in quiet mode) the warning message is also sent to the RS-232  
interface. These messages are helpful when trying to track down a problem with the analyzer and  
for determining whether or not the DAS reports are actually valid. The warning message format  
is for example:  
W 194:11:03 0200 SAMPLE FLOW WARN  
The format of a warning command is W command. Examples of warning commands are:  
W LIST  
List all current warnings  
Clear all current Warnings  
W CLEAR ALL  
Individual warnings may be cleared via the front panel or the command line interface. To clear  
the sample flow warning shown above the command would be:  
W WSAMPFLOW  
5.5.5 CALIBRATION Commands and Messages  
There are several methods of both checking the calibration and calibrating the M200AU, these  
are discussed in Section 7. The C command executes one of the calibration commands shown in  
Table 5-18.  
Table 5-18: RS-232 Calibration Commands  
Command  
Description  
C [id] ZERO [1 or 2]  
Start remote zero calibration. The number is optional and selects the  
range to calibrate. If not specified, the range defaults to range 1.  
C [id] COMPUTE ZERO  
Tells the instrument to compute a new slope and offset. Same as  
pressing ZERO-ENTR on front panel.  
C [id] SPAN [1 or 2]  
Start remote span calibration.  
C [id] COMPUTE SPAN  
Tells the instrument to compute a new slope and offset. Same as  
pressing SPAN -ENTR on front panel.  
C [id] ASEQ number  
C [id] EXIT  
Executes automatic calibration sequence (1, 2, or 3).  
Exits the current calibration step and goes to the next one.  
Aborts the entire calibration sequence.  
C [id] ABORT  
5-34  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 5-19: RS-232 Calibration Examples  
Action  
RS-232 Commands  
Comments  
Zero Calibration  
C ZERO  
C COMPUTE ZERO  
C EXIT  
Z/S valves switched to admit  
zero gas. Zero cal in Single  
Range mode  
Zero Calibration of low range -  
AutoRange Enabled  
C ZERO 1  
C COMPUTE ZERO  
C EXIT  
Z/S valves switched to admit  
zero gas. Zero calibration of  
low range in Auto Range mode  
Span Calibration of high range -  
AutoRange Enabled  
C SPAN 2  
C COMPUTE SPAN  
C EXIT  
Z/S valves switched to admit  
span gas. Span calibration of  
high range in Auto Range  
mode  
Zero Calibration with Dynamic  
Calibration enabled  
C ZERO  
C EXIT  
Z/S valves switched to admit  
zero gas. Instrument is zero  
calibrated if DYN CAL is  
enabled.  
Zero Calibration  
C ZERO  
C EXIT  
Z/S valves switched to admit  
zero gas. Instrument zero is  
just checked, but not changed.  
Execute AutoCal Sequence #2  
C ASEQ 2  
Execute a predefined AutoCal  
Sequence. Executes sequence  
immediately, ignoring time and  
date parameters.  
Span Calibration Check  
C SPAN  
C EXIT  
Z/S valves switched to admit  
span gas. Instrument span is  
just checked, but not changed.  
Whenever the analyzer starts or finishes a ZS calibration, it issues a status report to the RS-232  
interface. If the RS-232 interface is in the normal mode, these reports will be sent. Otherwise,  
they will be discarded. Table 5-20 shows the format of the text of the calibration messages. An  
example of an actual sequence of calibration status messages is:  
C DDD:HH:MM IIII START MULTI-POINT CALIBRATION  
C DDD:HH:MM IIII NOX=xxxxx PPB NO=xxxxx PPB NO2=xxxxx PPB  
C DDD:HH:MM IIII FINISH MULTI-POINT CALIBRATION  
5-35  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 5-20: RS-232 Calibration Messages  
Message  
Description  
START ZERO CALIBRATION  
NOX1=xxxxx PPB2 NO1=xxxxx PPB2 NO21=xxxxx PPB2  
FINISH ZERO CALIBRATION  
Beginning ZS zero calibration  
Finished ZS zero calibration  
START SPAN CALIBRATION  
NOX1=xxxxx PPB2 NO1=xxxxx PPB2 NO21=xxxxx PPB2  
FINISH SPAN CALIBRATION  
Beginning ZS span calibration  
Finished ZS span calibration  
START MULTI-POINT CALIBRATION  
NOX1=xxxxx PPB2 NO1=xxxxx PPB2 NO21=xxxxx PPB2  
FINISH MULTI-POINT CALIBRATION  
Beginning multi-point calibration  
Finished multi-point calibration  
1Depends on software options installed.  
2Depends on which units are currently selected.  
5.5.6 DIAGNOSTIC Commands and Messages  
When Diagnostic mode is entered from the RS-232 port, the diagnostic mode issues additional  
status messages to indicate which diagnostic test is currently selected. Examples of Diagnostic  
mode messages are:  
D DDD:HH:MM IIII ENTER DIAGNOSTIC MODE  
D DDD:HH:MM IIII EXIT DIAGNOSTIC MODE  
Example of turning on the Ozone Generator via the RS-232 port:  
D ENTER SIG  
D OZONE_GEN=ON  
D EXIT  
The following is a summary of the Diagnostic commands.  
5-36  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 5-21: RS-232 Diagnostic Command Summary  
Command  
Description  
Command  
Description  
D [id] LIST  
D [id] name=value  
Prints all I/O signal values. See Table 9-4 for Signal I/O definitions.  
Examines or sets I/O signal. For a list of signal names see Table 9-4.  
Must issue D ENTER SIG command before using this command.  
D [id] LIST NAMES  
Prints names of all diagnostic tests.  
D [id] ENTER SIG  
D [id] ENTER OT  
D [id] ENTER ET  
Executes SIGNAL I/O diagnostic test.  
Executes Optic Test diagnostic test.  
Executes Elect Test diagnostic test.  
Example of Ozone Generator diagnostic is in Section 5.5.6.  
Use D EXIT to leave these diagnostic modes.  
D [id] EXIT  
Must use this command to exit SIG, ET or OT Diagnostic modes  
Resets analyzer software (same as power on).  
D [id] RESET  
D [id] RESET RAM  
Resets analyzer software and erases RAM. Erases NO, NOx, NO2 conc  
values. Keeps setup variables and calibration. (same as installing new  
software version)  
D [id] RESET EEPROM  
Resets analyzer software and erases RAM and EEPROM. Returns all  
setup variables to factory defaults, resets calibration, Pre-reactor values.  
5.5.7 DAS Commands and Message  
The M200AU contains a flexible and powerful built in data acquisition system (DAS) that  
enables the analyzer to store concentration data as well as diagnostic parameters in its battery  
backed memory. This information can be printed out through the RS-232 port. The diagnostic  
data can be used for performing “Predictive Diagnostics” and trending to determine when  
maintenance and servicing will be required.  
To print out the properties of all of the data channels enter:  
D PRINT  
To print the properties of just a single data channel enter:  
D PRINT "name”  
For example to print the properties of the CONC data channel enter:  
D PRINT “CONC”  
5-37  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
To print records from a DAS data channel enter:  
D REPORT “name” RECORDS=nnn COMPACT|VERBOSE  
Examples of reports are:  
D REPORT “CONC” RECORDS=35 VERBOSE  
D REPORT “CALDAT” RECORDS=10  
D REPORT “PNUMTC” RECORDS=155 VERBOSE  
Automatic RS-232 reporting can be independently enabled and disabled for each Data Channel.  
For all default data channels, automatic reporting is initially set to “OFF.” If this property is  
turned on, the Data Channel will issue a report with a time and date stamp to the RS-232 port  
every time a data point is logged. The report format is shown below:  
D 94:08:00 0200 CONC : AVG NXCNC1 = 1234.5 PPB  
D 94:08:00 0200 CONC : AVG NOCNC1 = 1234.5 PPB  
D 94:08:00 0200 CONC : AVG N2CNC1 = 1234.5 PPB  
One CONC report consists of:  
D
= Type of report (Diagnostic)  
= Time and Date stamp (Julian day, Hr, Min)  
= Instrument ID number  
= Data Channel name  
94:08:00  
0200  
CONC  
CONC = concentration data  
PNUMTC = pneumatic parameters  
CALDAT = calibration parameters  
= Type of data  
AVG  
AVG = average reading  
INST = instantaneous reading  
= Name of the parameter  
NX = NOx  
NXCNC1 = 1234.5 PPB  
NO = NO  
N2 = NO2  
5-38  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
All of the default Data Channels sample more than one parameter, for these channels, each  
parameter is printed on a separate line.  
There is also a compact format. If this attribute is enabled, all 3 concentration parameters are  
printed on one line as shown below:  
D 94:08:00 0200 CONC : 120.0 100.0 20.0  
The parameters are in the order of NOx, NO, and NO2.  
To change any of the attributes of a particular data channel, the channel attributes are edited  
from the front panel by pressing the EDIT key.  
5.5.8 Internal Variables  
A list of M200AU variables is shown in Table 9-5.  
A list of variables and their settings can be requested over the RS-232 port by:  
V LIST  
Lists internal variables and values  
The output from this command is long and will not be shown here. The general format of the  
output is:  
name = value warning_lo warning_hi (data_lo to data_hi)  
Where:  
name  
= name of the variable  
value  
= current value of variable  
warning_lo  
warning_hi  
data_lo  
data_hi  
= lower limit warning (displayed if applicable)  
= upper limit warning (displayed if applicable)  
= lower limit of allowable values  
= upper limit of allowable values  
5-39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Variables can be changed. Before changing the settings on any variables, please make sure you  
understand the consequences of the change. We recommend you call the factory before changing  
the settings on any variables. The general format for changing the settings on a variable is:  
V name[=value [warn_lo [warn_hi]]]  
For example to change the warning limits on the box temperature type:  
V BOX_SET 30 10 50  
and the CPU should respond with:  
V DDD:HH:MM IIII BOX_SET=30 10 50 (0 to 60)  
The CONFIG command lists the software configuration. To show the software configuration,  
type:  
V CONFIG  
In addition to SAMPLE and SETUP modes the M200AU has a number of additional operational  
modes. They are listed in Table 5-8. To list the analyzer's current mode type:  
V MODE  
5-40  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
6 OPTIONAL HARDWARE AND SOFTWARE  
Optional equipment offered with the M200AU includes:  
1. Rack mount with slides  
2. Rack mount without slides, ears only  
3. Stainless steel zero/span valves  
4. 4-20mA, isolated outputs  
5. NOy Converter  
6.1 Rack Mount Options  
Rack Mount including slides and ears, permits the Analyzer to be mounted in a standard 19"  
wide x 24" deep RETMA rack. Can also be ordered without slides for applications requiring the  
instrument to be rigidly mounted in a RETMA rack.  
6.2 Zero/Span Valves  
The Zero/Span Valve option consists of two stainless steel solenoid valves. This option is  
applicable only to the instrument with internal molybdenum converter. See Figure 2-5 for valve  
location. Connections are provided on the rear panel for span gas and zero gas inputs to the  
valves. (See Figure 2-2) The valves can be actuated by several methods shown in Table 6-1.  
Table 6-1: Zero/Span Valve Operation  
Mode  
Description  
Reference Section  
1.  
Front panel operation via  
CALS and CALZ buttons  
Calibration Section 7 - Manual Zero/Span Check.  
2.  
3.  
Automatic operation using  
AUTOCAL  
Setup and use of AUTOCAL is described in Table 6-2,  
and Section 7.3.  
Remote operation using the  
RS-232 interface  
Setup described in Table 6-2. Operation of AUTOCAL  
described in Section 5.3.2 and Section 7 - Calibration. A  
complete description of the RS-232 interface is available.  
Order part number 01350.  
4.  
Remote operation using  
external contact closures  
Section 7.7 - Automatic operation using external contact  
closures. Truth Table 7-8 and Section 9.3.4.3.  
6-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Zero/Span valves have 3 operational states:  
1. Sample mode. Here both valves are un-energized and sample gas passes through the  
sample/cal valve and into the analyzer for analysis.  
2. Zero mode. The sample/cal valve is energized to the cal mode. The zero/span valve is un-  
energized in the zero mode, thus allowing zero gas to be admitted through the rear panel  
bulkhead fitting into the analyzer.  
3. Span mode. The sample/cal and the zero/span valves are both energized and in the cal mode.  
With both valves on, span gas is admitted through a rear panel bulkhead fitting into the  
analyzer.  
Zero air and span gas inlets should supply their respective gases in excess of the 1000 cc/min  
demand of the Analyzer. Supply and vent lines should be of sufficient length and diameter to  
prevent back diffusion and pressure effects. See Figure 2-3 for fitting location and tubing  
recommendations.  
Zero air for this instrument should have less than 1 ppt (.001 ppb) NO and have various  
interferent gas concentrations such that they produce less than 1 ppt total interferent. We  
recommend a tank of Ultra Zero Air commonly used in gas chromatographic applications as a  
reliable and constant source of zero air. Another source of zero air is the Model 701 zero air  
module.  
6.3 Autocal - Setup Zero/Span Valves  
The Autocal system operates by executing SEQUENCES. It is possible to enable 0-3 sequences,  
each sequence operates in one of 4 MODES:  
Mode No.  
Mode Name  
Disabled  
Zero  
Action  
1.  
2.  
3.  
4.  
Disables the Sequence  
Does a Zero Calibration  
Does a Zero and Span Calibration  
Does a Span Calibration  
Zero-Span  
Span  
6-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
For each mode there are seven attributes that the MODE can have that control operational details  
of the SEQUENCE. They are:  
Attribute No.  
Attribute Name  
Timer Enabled  
Starting Date  
Starting Time  
Delta Days  
Action  
1.  
2.  
3.  
4.  
5.  
6.  
7.  
Turns on the Sequence timer  
Sequence will operate after Starting Date  
Time of day sequence will run  
Number of days to skip between each Seq. execution  
Number of hours later each “Delta Days” Seq is to be run.  
Number of minutes the sequence operates  
Calibrate the instrument at end of sequence.  
Delta Time  
Duration  
Calibrate  
Example of enabling sequence #2:  
Do a span check ½ hour later every other day, lasting 15 minutes, without calibration.  
Mode and Attribute  
Sequence  
Value  
Comment  
2
Define Seq. #2  
Mode  
4
Select Span Mode  
Timer Enable  
Starting Date  
Starting Time  
Delta Days  
Delta Time  
Duration  
ON  
Enable the timer  
Sept. 4, 1996  
Start after Sept 4  
01:00  
2
First Span starts at 1:00AM  
Do Seq #2 every other day  
Do Seq #2 ½ hr later each time  
Operate Span valve for 15 min  
Do not calibrate at end of Seq  
00:30  
15.0  
NO  
Calibrate  
6-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 6-2: Example of AutoCal Setup  
Step Action  
Comment  
1.  
Press  
SETUP-ACAL  
This button sequence will cause the AUTOCAL menu to be  
displayed.  
2.  
3.  
3.  
4.  
5.  
6.  
7.  
8.  
9.  
Press PREV-NEXT  
Press MODE  
Press PREV-NEXT until SEQ 2 is displayed  
Select the MODE menu  
Press PREV-NEXT  
Press ENTR  
Press PREV NEXT to scroll to SPAN  
ENTR selects the SPAN MODE  
Press SET  
Select the SET menu to change the sequence attributes  
Scroll the SET menu to TIMER ENABLE  
Allows changing the TIMER ENABLE attribute, select ON  
ENTR changes TIMER ENABLE to ON  
Repeat steps 6-9 for each attribute  
Press PREV-NEXT  
Press EDIT  
Press ENTR  
Press PREV-NEXT  
10. Press EXIT  
Press the EXIT key to return to upper level menus  
6.4 4-20 mA, Current Loop Output  
The current loop option replaces the voltage output of the instrument with an isolated 4-20 mA  
current output. The current outputs come out on the same terminals that were used for voltage  
outputs, see Figure 2-2. See Troubleshooting Section 9.3.3 for setup and calibration.  
6.5 NOy Converter  
The NOy Converter Option allows placement of the molybdenum converter at the sample inlet  
point. This configuration is useful for background studies and where it is desirable to  
immediately convert certain components of the atmosphere that decompose rapidly, or may be  
lost on the walls of a normal sample induction system. This group of compounds is sometimes  
referred to as NOy.  
The Option consists of a modified M200AU, a M501Y chassis and remote converter. The  
pneumatic diagram for this system of components is in Figure 8-7. Because of the nature of this  
option, there is a separate manual “M501Y – Remotely Mounted Converter” (P/N 02808). The  
manual covers setup, calibration, and operation of the system.  
6-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
7 CALIBRATION & ZERO/SPAN CHECKS  
There are several ways to check and adjust the calibration of the M200AU. These different  
methods are summarized in Table 7-1. In addition, all of the methods described in this section  
can be initiated and controlled via the RS-232 port.  
We strongly recommend that SPAN CALIBRATION be done with NO span gas. SPAN  
CHECKS can be done with either NO only, NO2 only or a mixture of NO and NO2 (GPT).  
Zero air used for all calibration procedures, including GPT, should have 1 ppt NO and NO2, less  
than 1 ppt of major interferents such as SO2, NH3, and hydrocarbons and a dew point of -5o C or  
less. The calibration gasses should be from a reliable supplier, since the quality of the tank  
concentration values ultimately determines the accuracy of the analyzer. EPA protocol  
calibration gasses should be used for EPA monitoring, see Section 7.6.  
NOTE  
If you are using the M200AU for EPA monitoring, only the  
calibration method described in Section 7.6 should be used.  
NOTE  
If there are any problems completing any of the following procedures,  
refer to Section 9.2.8 and 9.2.9 - Unable to Span or Zero.  
7-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 7-1: Types of Zero/Span Check and Calibration  
Section  
Type of Cal or Check  
Description  
7.1  
Manual Z/S Check or Calibration  
through the sample port  
This calibration option uses calibration gas  
coming in through the sample port. Zero/Span  
valves do not operate.  
7.2  
7.3  
7.4  
7.5  
Manual Z/S Check or Calibration with How to operate Zero/Span Valves Option. Can be  
Z/S Valves Option. used to check or adjust calibration.  
Automatic Z/S Check with Z/S Valves Operates Z/S valves once per day to check the  
calibration.  
Dynamic Z/S Calibration with Z/S  
Valves  
Operates Z/S valves once per day and adjusts  
calibration.  
Use of Z/S Valves with Remote  
Contact Closure  
Operates Z/S valves with rear panel contact  
closures. Without valves, can be used to switch  
instrument into zero or span cal mode. Used for  
either checking or adjusting zero/span.  
7.6  
7.7  
7.8  
EPA Protocol Calibration  
Covers methods to be used if data is for EPA  
equivalency monitoring.  
Special Calibration Requirements for  
Independent Ranges or AutoRanging  
Covers special requirements if using Independent  
Range or AutoRange  
Calibration Quality  
Information on how to determine if the calibration  
performed will result in optimum instrument  
performance.  
7.9  
References  
Contains a list of references on quality control  
and calibration.  
7-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Figure 7-1: Calibration Setup  
7-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
7.1 Manual Zero/Span Check or Cal with Zero/Span Gas in  
the Sample Port  
The calibration of the instrument can be checked or adjusted using gas coming through the  
sample port. This method is often used when the calibration gas is supplied from an external  
calibrator and valve system.  
This is the calibration procedure to use if the instrument is purchased without the Zero/Span  
Valve option.  
Since the zero gas concentration is defined as 0 ppb, it is not necessary to enter the expected zero  
value. Table 7-2 details the zero calibration procedure with zero gas coming in through the  
sample port.  
Table 7-2: Manual Zero Calibration Procedure - Zero Gas thru Sample Port  
Step Number  
Action  
Comment  
1.  
Press CAL  
The M200AU enters the calibrate mode from sample mode.  
The zero gas must come in through the sample port. When the  
CAL button is pressed, the adaptive filter is activated. This  
allows the instrument to respond rapidly to concentration  
changes regardless of their magnitude.  
2.  
3.  
Wait 10 min  
Press ZERO  
Wait for reading to stabilize at zero value or for STABIL  
reading to get to < 2 ppb.  
If you change your mind after pressing ZERO, you can still  
press EXIT here without zeroing the instrument.  
4.  
5.  
Press ENTR  
Press EXIT  
Pressing ENTR actually changes the calculation equations.  
M200AU returns to sampling. Immediately after calibration,  
data is not added to the DAS averages.  
7-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Enter the expected NOx and NO span gas concentrations:  
Table 7-3: Enter Expected Span Gas Concentrations Procedure  
Step Number  
Action  
Comment  
1.  
Press  
This key sequence causes the M200AU to prompt for the  
CAL-CONC-NOX expected NOx concentration.  
Enter the NOx span concentration value by pressing the key  
under each digit until the expected value is set. This menu can  
also be entered from CALS or CALZ.  
2.  
3.  
Press ENTR  
ENTR stores the expected NOx span value.  
Press  
Now enter the expected NO span concentration as in step one.  
CAL-CONC-NO  
4.  
5.  
Press ENTR  
Press EXIT  
Pressing ENTR stores the NO span value and returns the  
prompt to the CONC menu.  
Returns instrument to SAMPLE mode.  
If desired, compensation for moly converter efficiency (CE) can be included in the NOx  
concentration calculation. The CE must be entered prior to calibration. Refer to Section 7.6.6.1  
for the CE procedure.  
Table 7-4: Manual Span Calibration Procedure - Span Gas thru Sample Port  
Step Number  
Action  
Comment  
1.  
Press CAL  
The M200AU enters the calibrate mode. NO span gas should  
be fed to the sample port. When the CAL button is pressed,  
the adaptive filter is activated. This allows the instrument to  
respond rapidly to concentration changes regardless of their  
magnitude.  
2.  
3.  
Wait 10 min  
Press SPAN  
Wait for reading to stabilize at span value.  
If you change your mind after pressing SPAN, you can still  
press EXIT here without spanning the instrument.  
4.  
5.  
Press ENTR  
Press EXIT  
Pressing ENTR actually changes the calculation equations and  
causes the instrument to read the NO and NOx span  
concentrations.  
M200AU returns to sampling. Immediately after calibration,  
data is not added to the DAS averages.  
7-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
7.2 Manual Zero/Span Check or Calibration with  
Zero/Span Valves Option  
The Zero/Span valve option can be operated from the front panel keyboard. In the Zero/Span  
valve option the zero and span gas come into the valves through ports on the rear panel of the  
instrument.  
Table 7-5: Manual Zero Calibration Procedure - Z/S Valves  
Step Number  
Action  
Comment  
1.  
Press CALZ  
The analyzer enters the zero calibrate mode. This switches the  
sample/cal and zero/span valves to allow zero gas to come in  
through the zero gas inlet port in the rear panel. When the  
CALZ button is pressed, the adaptive filter is activated. This  
allows the instrument to respond rapidly to concentration  
changes regardless of their magnitude.  
2.  
3.  
Wait 10 min  
Press ZERO  
Wait for reading to stabilize at zero value.  
If you change your mind after pressing ZERO, you can still  
press EXIT here without zeroing the instrument.  
4.  
5.  
Press ENTR  
Press EXIT  
Pressing ENTR actually changes the calculation equations,  
forcing the reading to zero.  
M200AU returns to sample mode. Immediately after  
calibration, readings do not go into the DAS averages.  
Refer to Table 7-3 to enter expected NO and NOx values.  
7-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 7-6: Manual Span Calibration Procedure - Z/S Valves  
Step Number  
Action  
Comment  
1.  
Press CALS  
The M200AU enters the calibrate mode from sample mode.  
This operates the sample/cal and zero/span valves to allow  
span gas to come in through the cal gas inlet port in the rear  
panel. When the CALS button is pressed, the adaptive filter is  
activated. This allows the instrument to respond rapidly to  
concentration changes regardless of their magnitude.  
2.  
3.  
Wait 10 min  
Press SPAN  
Wait for reading to stabilize at span value.  
If you change your mind after pressing SPAN, you can still  
press EXIT here without spanning the instrument.  
4.  
5.  
Press ENTR  
Press EXIT  
Pressing ENTR actually changes the calculation equations.  
M200AU returns to sampling. Immediately after calibration,  
data is not added to the DAS averages.  
7.3 Automatic Zero/Span Check  
In a typical air monitoring application it is desirable to have the analyzer automatically check  
(AUTOCAL) its calibration each day. If provided with the proper options, the M200AU  
provides this capability by using the time of day clock to signal the computer system to check  
operations. When enabled, the instrument software will automatically check zero and span  
(AUTOCAL) each day. Optionally, the Z/S cycle can be moved backwards or forwards a fixed  
time each day.  
AUTOCAL setup is covered in Table 6-2.  
7.4 Dynamic Zero/Span Calibration  
The AUTOCAL system described above can also optionally be used to calibrate the instrument  
once each 24 hours. Dynamic Calibration is enabled by the CALIBRATE attribute in the  
AUTOCAL setup menu, see Table 6-2 and Table 7-7. If calibration is initiated using the rear  
panel contact closures, the DYN_ZERO and DYN_SPAN variables must be set to ON in the  
VARS menu.  
Before proceeding with enabling DYNAMIC Z/S you must setup the AUTOCAL feature.  
Enabling AUTOCAL is described in Table 6-2. To enable DYNAMIC Zero/Span Calibration:  
7-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 7-7: Enabling Dynamic Zero/Span  
Step Number  
Action  
Comment  
1.  
Press  
Causes the M200AU go to the AutoCal menu.  
SETUP-ACAL  
2.  
3.  
4.  
5.  
Press PREV-NEXT Select the sequence you want for dynamic calibration.  
Press SET Select the SET menu.  
Press PREV-NEXT Scroll through the SET menu to the CALIBRATE attribute.  
Press EDIT  
Set the CALIBRATE attribute value to ON to enable  
Dynamic Span.  
6.  
Press EXIT  
Causes the M200AU to return to SAMPLE mode.  
With dynamic calibration turned on, the instrument will re-set the slope and offset values for the  
NO and NOx channel each day. This continual re-adjustment of calibration parameters can often  
mask subtle fault conditions in the analyzer. It is recommended that if Dynamic Cal is enabled,  
the TEST functions, SLOPE and OFFSET values in the M200AU should be checked frequently  
to assure high quality and accurate data from the instrument.  
7.5 Use of Zero/Span Valves with Remote Contact Closure  
The Zero/Span valve option can be operated using Remote Contact Closures provided on the rear  
panel. See Figure 2-2 for connector location and pinout. When the contacts are closed, the  
analyzer will switch to zero or span mode. The contacts must remain closed for at least 1 second,  
and will remain in zero or span mode as long as the contacts are closed. To calibrate the  
instrument at the end of the Zero/Span valve cycle (DYNAMIC CAL), the CAL_ZERO and  
CAL_SPAN variables in the VARS menu must be set to ON.  
The CPU monitors these two contact closures and will switch the Analyzer into zero or span  
mode when the contacts are closed for at least 1 second.  
In order to do another remote check, both contact closures should be held open for at least 1  
second, then may be set again. Table 7-8 shows what type of check is performed based on the  
settings of the two contact closures.  
7-8  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 7-8: Z/S Valves Modes with Remote Contact Closure  
Ext Zero CC  
Contact Open  
Contact Open  
Contact Closed  
Ext Span CC  
Contact Open  
Contact Closed  
Contact Open  
Operation  
State when in SAMPLE mode, normal monitoring.  
Span check or calibrate*  
Zero check or calibrate*  
* Calibrate only if Dynamic Calibration is enabled.  
7.6 EPA Protocol Calibration  
If the M200AU is to be used for EPA compliance monitoring, it must be calibrated in accordance  
with the instructions in this section.  
In order to insure that high quality, accurate measurements are obtained at all times, the  
M200AU must be calibrated prior to use. A quality assurance program centered on this aspect  
and including attention to the built-in warning features of the M200AU, periodic inspection,  
regular zero/span checks and routine maintenance is paramount to achieving this.  
In order to have a better understanding of the factors involved in assuring continuous and  
reliable information from the M200AU, it is strongly recommended that Publication No. PB 273-  
518 Quality Assurance Handbook for Air Pollution Measurement Systems (abbreviated, Q.A.  
Handbook) be purchased from the NTIS (phone 703-487-4650). Special attention should be paid  
to Section 2.3 which deals with chemiluminescent based NO2 analyzers and upon which most of  
this section is based. Specific regulations regarding the use and operation of ambient oxides of  
nitrogen analyzers can be found in 40 CFR 50 and 40 CFR 58. Both publications are available  
from the U.S. Government Printing Office (phone 202-783-3238).  
7.6.1 Calibration of Equipment  
In general, calibration is the process of adjusting the gain and offset of the M200AU against  
some recognized standard. The reliability and usefulness of all data derived from any analyzer  
depends primarily upon its state of calibration. In this section the term dynamic calibration is  
used to express a multipoint check against known standards and involves introducing gas  
samples of known concentration into the instrument in order to adjust the instrument to a  
predetermined sensitivity and to produce a calibration relationship. This relationship is derived  
from the instrumental response to successive samples of different known concentrations. As a  
minimum, three reference points and a zero point are recommended to define this relationship.  
The true values of the calibration gas must be traceable to NIST-SRM's (Section 2.0.7, Q.A.  
Handbook).  
7-9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
All monitoring instrument systems are subject to some drift and variation in internal parameters  
and cannot be expected to maintain accurate calibration over long periods of time. Therefore, it  
is necessary to dynamically check the calibration relationship on a predetermined schedule. Zero  
and span checks must be used to document that the data remains within control limits. These  
checks are also used in data reduction and validation. Table 7-9 summarizes the initial quality  
assurance activities for calibrating equipment. Table 7-10 is a matrix for the actual dynamic  
calibration procedure.  
Table 7-9: Activity Matrix for Calibration Equipment and Supplies  
Equipment/  
Supplies  
Frequency And Method Action If Requirements  
Acceptance Limits  
Of Measurement  
Are Not Met  
Recorder  
Compatible with output  
signal of analyzer; min  
chart width of 150 mm  
(6 in) is recommended  
Check upon receipt  
Return equimpent to  
supplier  
Sample line  
and manifold  
Constructed of PTFE,  
glass, or stainless steel  
Check upon receipt  
Return equipment to  
supplier  
Calibration  
equipment  
Meets guide line reference See Section 2.3.9 (Q. A.  
Return equipment/ supplies  
to supplier or take corrective  
action  
1 and Section 2.3.2(Q. A.  
Handbook)  
Handbook)  
Working  
standard NO  
cylinder gas or protocol for accuracy and  
Traceable to NIST-SRM  
Meets limits in traceability SRM; see protocol in  
Analyzed against NIST-  
Obtain new working  
standard and check for  
traceability  
Section 2.0.7, Q.A.  
NO2  
stability. (Section 2.0.7, Q. Handbook  
permeation  
tube  
A. Handbook)  
Recording  
forms  
Develop standard forms  
N/A  
Revise forms as appropriate  
Audit  
equipment  
Must not be the same as  
used for calibration  
System must be checked  
out against known  
standards  
Locate problem and correct  
or return to supplier  
Calibrations should be carried out at the field monitoring site. The Analyzer should be in  
operation for at least several hours (preferably overnight) before calibration so that it is fully  
warmed up and its operation has stabilized. During the calibration, the M200AU should be in the  
CAL mode, and therefore sample the test atmosphere through all components used during  
normal ambient sampling and through as much of the ambient air inlet system as is practicable.  
If the instrument will be used on more than one range, it should be calibrated separately on each  
applicable range (See Section 7.7). Calibration documentation should be maintained with each  
analyzer and also in a central backup file.  
7-10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 7-10: Activity Matrix for Calibration Procedure  
Equipment/  
Supplies  
Frequency And Method Action If Requirements  
Acceptance Limits  
Of Measurement  
Are Not Met  
Calibration  
gases  
Sec. 2.0.7, Subsec. 7.1  
(Q.A. Handbook)  
Assayed against an  
NIST-SRM quarterly,  
Sec. 2.0.7, (Q.A.  
Handbook)  
Working gas standard is  
unstable, and/or  
measurement method is  
out of control; take  
corrective action such as  
obtaining new calibration  
gas.  
Dilution gas  
Zero air, free of  
See TAD2  
Return to supplier or take  
appropriate action with  
generation system  
contaminants; TAD2 and  
Sec. 2.0.7, Subsec. 7.1  
(Q.A. Handbook)  
Multi-point  
calibration  
(GPT)  
1. tR < 2 minutes PR >  
2.75 ppm/min  
Method  
1. Adjust flow conditions  
and/or reaction chamber  
volume to meet  
1. Sect. 7.6.4 (this  
manual)  
2. Use calibration  
procedure in  
suggested limits  
2. Section 7.6.5.3 (this  
manual), TAD2,  
Federal Register and  
Figure 7-2; see  
Subsection 2.4 (Q.A.  
Handbook); also TAD2  
and Federal Register  
2. Repeat the calibration  
3. Replace or service the  
converter  
3. Converter efficiency >  
96%  
Section 7.6.7 for  
frequency  
3. Subsection 7.6.6 (this  
manual)  
7.6.2 Calibration Gas and Zero Air Sources  
Production of Zero Air  
Due to the high sensitivity of the M200AU special care must be taken to assure that the zero air  
has extremely low levels of pollutants. We recommend using Ultra Zero Air, which is commonly  
used for gas chromatographic applications. If other zero air sources are used, the NO  
concentration should be < 1 ppt (0.001 ppb) and the concentration of interferent gasses should be  
< 1 ppt (0.001 ppb) for the sum total of all interferents.  
Devices that condition ambient air by drying and removal of pollutants are available on the  
commercial market such as the Teledyne API Model 701 Zero Air Module. We recommend this  
type of device for generating zero air. Detailed procedures for generating zero air are in TAD2.  
7-11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Selection of NO span gas standards  
The NIST-SRM's provide references against which all calibration gas mixtures must be  
compared (Section 2.0.7, Q.A. Handbook). The procedure requires the comparison of the  
concentration of a commercial, working calibration standard to an NIST-SRM. This is described  
in Subsection 7.1 of Section 2.0.7, Q.A. Handbook. Subsections 7.1.4 and 7.1.5 describe the  
verification and re-analysis of cylinder gases.  
Care must be taken to assure that no oxygen is allowed in the NO tank, regulator, or sample  
lines. If oxygen is present, it reacts with NO to produce NO2, which can lead to significant errors  
in NO concentration measurement.  
It is good practice to request a NOx analysis from the supplier of the NO calibration gas.  
Generally the maximum NO2 impurity that should be allowed is 1% of the NO concentration.  
The NO2 impurity in the NO standard can be measured by the M200AU. A procedure is given in  
the TAD for NO2 measurement.  
Use of NO2 permeation tubes as standards  
The steps required to compare the concentration of a commercial working calibration standard to  
an NIST-SRM are described in Subsection 7.3.3 of Section 2.0.7, Q.A. Handbook. See  
Subsection 7.3.6 for the re-analysis of permeation tubes.  
7.6.3 Data Recording Device  
Either a strip chart recorder, data acquisition system, digital data acquisition system should be  
used to record the data from the M200AU RS-232 port or analog outputs. If analog readings are  
being used, the response of that system should be checked against a NIST referenced voltage  
source or meter. Data recording device should be capable of bi-polar operation so that negative  
readings can be recorded.  
7.6.4 Gas Phase Titration (GPT) System  
7.6.4.1 Gas Phase Titration (GPT  
Gas Phase Titration (GPT) with serial dilution is recommended for checking the converter  
efficiency and NO2 channel linearity of the M200AU. Those using a NO2 permeation tube  
should refer to TAD.2  
The principle of GPT is based on the rapid gas phase reaction between NO and O3 which  
produces stoichiometric quantities of NO2 as shown by the following equation:  
NO + O3 → NO2 + O2  
7-12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Given that the NO concentration is known for this reaction, the resultant concentration of NO2  
can be determined. Ozone is added to excess NO in a dynamic calibration system, and the NO  
channel of the chemiluminescent analyzer detects the changes in NO concentration. After the  
addition of O3, the observed decrease in NO concentration on the calibrated NO channel is  
equivalent to the concentration of NO2 produced. The amount of NO2 generated may be varied  
by adding varying amounts of O3 from a stable O3 generator. All zero air used in this procedure  
should conform to the requirements stated in Section 7.  
Dynamic calibration systems based on this principle are commercially available such as the  
Teledyne API Model 700 Calibrator, or may be assembled by the user. A recommended  
calibration system is described in the Federal Register1 and detailed in TAD.2  
7.6.4.2 GPT Calibrator Check Procedure  
It has been determined empirically that the NO-O3 reaction goes to completion (<1% residual  
O3) if the NO concentration in the reaction chamber (ppm) multiplied by the residence time  
(min.) of the reactants in the chamber is >2.75 ppm-min. The theory behind the development of  
this equation is in the Federal Register1 and in TAD.2  
It is currently not known whether this relationship holds up at the extremely low concentrations  
of which the M200AU is capable. We therefore recommend that the GPT procedure be carried  
out at the normal flow-rate and concentrations encountered in ambient air monitoring, then  
performing a serial dilution of the resultant gas stream to get the low concentration values.  
The following procedures and equations should be used to determine whether an existing GPT  
calibration system will meet required conditions for a specific calibration.  
For calibrators that have known pre-set flow rates, use equations 7-5 and 7-6 of steps 7 and 8  
(below) to verify the required conditions. If the calibrator does not meet specifications, follow  
the complete procedure to determine what flow modifications must be made.  
1.  
2.  
3.  
Select a NO standard gas that has a nominal concentration in the range of 50 to 100 ppm.  
Determine the exact concentration [NO]STD by referencing against an NIST-SRM, as  
discussed in Section 2.0.7 (Q.A. Handbook).  
Determine the volume (cm3) of the calibrator reaction chamber (VRC). If the actual  
volume is not known, estimate the volume by measuring the approximate dimensions of the  
chamber and using an appropriate formula.  
Determine the required minimum total flow output (FT) using Equation 7-1:  
FT = analyzer flow demand (cm3/min) x 110/100 Equation 7-1  
If more than one analyzer is to be calibrated at the same time, multiply FT by the number of  
analyzers.  
7-13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
4. Calculate the NO concentration [NO]OUT needed to approximate 90% of the URL of the NO2  
analyzer to be calibrated, using Equation 7-2:  
[NO]OUT = URL of analyzer (ppm) x 90/100 Equation 7-2  
5.  
6.  
Calculate the NO flow (FNO) required to generate the NO concentration [NO]OUT, using  
Equation 7-3:  
[NO  
x
]
FT  
OUT  
=
Equation 7-3  
FNO  
[NO  
]
STD  
Calculate the required flow through the ozone generator (FO), using Equation 7-4:  
[NO  
X
X
]
FNO V RC  
STD  
=
-
Equation 7-4  
Fo  
FNO  
2.75 ppm - min  
7.  
8.  
Verify that the residence time (tR) in the reaction chamber is <2 min, using Equation 7-5:  
VRC  
t
R
=
2min Equation 7-5  
FO  
+ FNO  
Verify that the dynamic parameter specification (PR) of the calibrator's reaction chamber  
is >2.75 ppm-min using Equation 7-6:  
FNO  
VRC  
PR  
=
[
NO  
]
STD  
×
×
2.75 Equation 7-6  
FO  
+ FNO  
FO  
+ FNO  
NOTE  
If tr is >2 minutes or if PR is <2.75 ppm-min, changes in flow conditions  
(FT, FO, FNO) or in the reaction chamber volume (VRC), or both will  
have to be made, and tr and PR will have to be re-calculated.  
9.  
After equations 7-5 and 7-6 are satisfied, calculate the diluent air flow (FD) using  
Equation 7-7:  
FD = FT - FO - FNO Equation 7-7  
7-14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
7.6.4.3 Example Calculation  
Following is an example calculation that can be used to determine whether an existing GPT  
calibrator will meet the required conditions for a specific calibration. For this example, it is  
assumed that only the volume of the reaction chamber, VRC, and the concentration of the NO  
standard, [NO]STD, are known. All flow settings (FNO, FO, FT, and FD) will be calculated. In many  
uses, these flow settings are known and need only to be substituted in Equations 7-5 and 7-6 to  
verify the required conditions. Before doing any calculations, the URL and flow demand of the  
analyzer being calibrated must be known. For the M200AU:  
Upper range limit = 0.5 ppm  
Flow demand = 1000 cm3/min.  
Volume of calibrator reaction chamber is determined by physical measurement:  
V
RC = 180 cm3  
The concentration of the NO standard gas to be used is determined by reference against an  
NIST-SRM (Section 2.0.7, Q.A. Handbook):  
[NO]STD = 50.5 ppm  
1.  
Determine the total flow (FT) required at the output manifold using Equation 7-1:  
FT = 1000 cm3/min (110/100) = 1100 cm3/min  
Because low flows are difficult to control and measure, it is often advantageous to set a  
higher total flow than needed. In this example, we will let FT = 3300 cm3/min  
2.  
3.  
Determine the NO conc, [NO]OUT, required at the output manifold, using Equation 7-2:  
[NO]OUT = 0.5 ppm (90/100) = 0.45 ppm  
Calculate the NO flow (FNO) required to generate [NO]OUT, using Equation 7-3:  
0.45 ppm× 3300 3 / min  
cm  
=
=
FNO  
50.5 ppm  
4.  
Calculate the required flow rate through ozone generator (FO) using Equation 7-4:  
3
cm  
50.5 ppm x 29.4 3 / min x 180  
cm  
=
- 29.4 3 / min  
cm  
FO  
=
2.75 ppm - min  
cm6 min  
2
97180  
/
- 29.4  
3 / min = 282  
cm  
3 / min  
cm  
7-15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
5.  
Verify that the residence time (tR) in the reaction chamber is <2 min using Equation 7-5:  
3
180  
cm  
t
R
=
= 0.58 min  
282 3 / min+ 29.5 3 / min  
cm cm  
6.  
Verify the dynamic parameter specification (PR) of the calibrator reaction chamber using  
Equation 7-6:  
3
29.4 3 / min  
cm  
180  
cm  
P
R
= 50.5 ppm×  
×
= 2.75 ppm  
282 3 / min+ 29.4 3 / min 282 3 / min+ 29.4 3 / min  
cm cm cm cm  
7.  
Calculate the diluent air flow (FD) required at the mixing chamber, using Equation 7-7:  
FD = 3300 cm3/min - 282 cm3/min –29.4 cm3/min = 2988.6 cm3/min  
7.6.5 Dynamic Multipoint Calibration Procedure  
The procedure for calibration of chemiluminescent NOx analyzers by GPT is specified in the  
Federal Register.1 This section applies the general procedure to the specific case of the  
M200AU.  
Calibration must be performed with a calibrator that meets all conditions specified in Subsection  
2.3.2 (Q.A. Handbook). Flow settings used in the GPT calibration for NO2 must be determined  
as illustrated in Section 7.6.4, this manual.  
The user should be sure that all flow meters are calibrated under the conditions of use against a  
reliable standard. All volumetric flow rates should be corrected to 25oC (78oF) and 760 mm  
(29.92 in.) Hg. Calibrations of flow meters are discussed in TAD.2  
Gas Phase Titration (GPT) requires the use of the NO channel of the analyzer to determine the  
amount of NO2 generated by titration. Therefore it is necessary to calibrate and determine the  
linearity of the NO channel before proceeding with the NO2 calibration. It is also necessary to  
calibrate the NOx channel. This can be done simultaneously with the NO calibration. During the  
calibration the M200AU should be operating in its normal sampling mode, and the test  
atmosphere should pass through all filters, scrubbers, conditioners, and other components used  
during normal ambient sampling and as much of the ambient air inlet system as is practicable.  
All operational adjustments to the M200AU should be completed prior to the calibration. The  
following software features must be set into the desired state before calibration.  
7-16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
1.  
2.  
3.  
4.  
Automatic Converter Efficiency compensation. See Sections 7.6.6, this manual.  
Independent range selection. See Sections 5.3.4 and 7.7, this manual.  
Automatic temperature/pressure compensation. See Table 9-5.  
Alternate units, make sure ppb units are selected for EPA monitoring. See Section  
5.3.4.5.  
5.  
Autoranging option. See Section 5.3.4.2.  
Converter efficiency should be set prior to calibration since its value is used in the computation  
of the NOx and NO2 concentration outputs.  
The analyzer should be calibrated on the same range used for monitoring.  
If AutoRanging or Independent range options are selected the highest of the ranges will result in  
the most accurate calibration, and should be used.  
Make sure the GPT calibration system can supply the range of concentrations at a sufficient flow  
over the whole range of concentrations that will be encountered during calibration.  
7-17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Figure 7-2: Diagram of GPT Calibration System  
7-18  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
7.6.5.1 Zero Calibration Procedure  
Since the zero gas concentration is defined as 0 ppb, it is not necessary to enter the expected zero  
value. The following Table 7-11 details the zero calibration procedure.  
Table 7-11: Zero Calibration Procedure  
Step Number Action  
Comment  
1.  
Press CAL  
The M200AU enters the calibrate mode from sample mode.  
NOTE:  
The analyzer does not operate the zero/span valves in this mode,  
the zero gas enters through the sample port. When the CAL  
button is pressed, the adaptive filter is activated. This allows the  
instrument to respond rapidly to concentration changes  
regardless of their magnitude.  
2.  
3.  
Wait 10 min  
Press ZERO  
Wait for reading to stabilize at the zero value.  
If you change your mind after pressing ZERO, you can still  
press EXIT here without zeroing the instrument.  
4.  
5.  
Press ENTR  
Press EXIT  
Pressing ENTR actually changes the calculation equations.  
M200AU returns to the SAMPLE mode.  
7.6.5.2 NO/NOx Calibration Procedure  
Adjust the NO concentration to approximately 80% of the URL of the NO channel. The expected  
NO and NOx span concentrations can be determined by measuring the cylinder and diluent flows  
and computing the resulting concentrations. If there is any NO2 impurity in the NO standard gas  
it should be taken into account when the NOx concentration is entered during the NO/NOx  
channel calibration. This is done by ADDING the impurity concentration to the NO  
concentration to get the NOx concentration for calibration. Calculate the exact NO and NOx  
concentrations as follows:  
]
x [NO  
FT  
FNO  
STD  
]
[NO  
=
Equation 7 - 8  
OUT  
Enter the respective concentrations using the procedure in Table 7-12. The expected span  
concentrations need not be re-entered each time a calibration is performed unless they are  
changed.  
Enter the expected NOx and NO span gas concentrations:  
7-19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 7-12: Expected Span Gas Concentration Procedure  
Step Number Action  
Comment  
1.  
Press  
This key sequence causes the M200AU to prompt for the  
CAL-CONC-NOX expected NOx concentration.  
Enter the NOx span concentration value by pressing the key  
under each digit until the expected value is set. This menu can  
also be entered from CALS or CALZ.  
2.  
3.  
Press ENTR  
ENTR stores the expected NOx span value.  
Press  
Now enter the expected NO span concentration as in step one.  
CAL-CONC-NO  
4.  
5.  
Press ENTR  
Pressing ENTR stores the NO span value and returns the prompt  
to the CONC menu.  
Press EXIT  
Returns instrument to SAMPLE mode.  
Sample the generated concentration until the NO and the NOx responses have stabilized.  
Span the instrument by the following procedure:  
Table 7-13: Span Calibration Procedure  
Step Number Action  
Comment  
1.  
Press CAL  
The M200AU enters the calibrate mode from sample mode.  
When the CAL button is pressed, the adaptive filter is activated.  
This allows the instrument to respond rapidly to concentration  
changes regardless of their magnitude.  
2.  
3.  
Wait 10 min  
Press SPAN  
Wait for readings to stabilize at span values.  
If you change your mind after pressing SPAN, you can still  
press EXIT here without spanning the instrument.  
4.  
5.  
Press ENTR  
Press EXIT  
Pressing ENTR actually changes the calculation equations.  
M200AU returns to SAMPLE mode.  
7-20  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
The analog voltage output should measure 80% of the voltage range selected. (e.g. 4.00VDC if  
0-5V output is selected.) The readings on the front panel display should be equal to the expected  
NO and NOx concentrations entered in the procedure given in Table 7-13. See the  
Troubleshooting Section 9.2.8 if there are problems. Also see the Calibration Quality Check  
procedure Section 7.8.  
After the zero and the 80% URL points have been set, generate five approximately evenly  
spaced calibration points between zero and 80% URL without further adjustment to the  
instrument. Allow the instrument to sample these intermediate concentrations for about 10  
minutes each and record the instrument NO and NOx responses.  
Plot the analyzer NO and NOx responses versus the corresponding calculated concentrations to  
obtain the calibration relationships. Determine the straight line of best fit (y = mx + b)  
determined by the method of least squares.  
After the best-fit line has been drawn for the NO and the NOx calibrations, determine whether  
the analyzer response is linear. To be considered linear, no calibration point should differ from  
the best-fit line by more than 2% of full scale.  
7.6.5.3 GPT - NO2 Calibration Procedure  
The M200AU computes the NO2 concentration by subtracting the NO from the NOx  
concentration. Unlike analog instruments, this difference is calculated by the M200AU's internal  
computer software. It is extremely unlikely that the NO2 concentration will be in error if the NO  
and NOx channels are calibrated correctly. Therefore this procedure is a confirmation that the  
NO2 subtraction algorithm in the computer is operating correctly.  
NOTE  
During this procedure do not make any adjustments to the instrument.  
1.  
Generate a NO concentration near 90% of the URL. Dilution air and O3 generator air  
flows should be the same as used in the calculation of specified conditions of the dynamic  
parameter according to Section 7.6.4. Sample this NO concentration until the NO and NOx  
responses stabilize. Record the NO and NOx concentrations.  
7-21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
2.  
Turn on and adjust the O3 generator in the calibrator to produce sufficient O3 to decrease  
the NO concentration to about 10% of full scale. This will be equivalent to 80% of the URL  
of the NO2 channel. After the analyzer responses stabilize, record the resultant NO, NOx, and  
NO2 concentrations.  
MOLY CONVERTER EFFICIENCY  
IF THE NOX READING SHOULD DROP TO LESS THAN 96% OF ITS STARTING  
VALUE DURING THIS STEP, IT INDICATES THE MOLY CONVERTER IS IN NEED  
OF TROUBLESHOOTING OR REPLACEMENT. SEE SECTION 7.6.6 FOR FURTHER  
DETAILS.  
3.  
4.  
While maintaining all other conditions, adjust the ozone generator to obtain several other  
concentrations of NO2 evenly spaced between the 80% URL point and the zero point. Record  
the NO, NOx, and NO2 concentrations for each additional point.  
Calculate the resulting NO2 concentrations as follows:  
]
* [  
FNO NO2  
IMP  
]
]
]
REM  
[
= [NO  
- [NO  
+
Equation 7 - 10  
NO2  
OUT  
ORIG  
FT  
Where:  
[NO]ORIG is the NO concentration before the GPT ozone is turned on, and [NO]REM is the NO  
remaining after GPT.  
5.  
Plot the NO2 concentration output by the instrument on the y-axis against the generated  
NO2 [NO2]OUT on the x-axis. The plot should be a straight line within the ± 2% linearity  
criteria given for the NOx and NO channels. If the plot is not linear the most likely cause is  
that the converter needs replacing. See Section 7.6.6 on Moly converter efficiency.  
7-22  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
7.6.6 Moly Converter Efficiency  
The moly efficiency should be 96 to 102% efficient. If it is outside these limits it should be  
replaced. The converter efficiency can be determined from the data collected in Section 7.6.5.3.  
For each NO2 concentration generated:  
1.  
Calculate the concentration of NO2 converted as:  
]
]
]
])  
NOx  
REM  
[
= [  
- ([  
- [  
Equation 7 - 11  
NO2  
NO2  
NOx  
CONV  
OUT  
ORIG  
Where:  
[NOx]ORIG is the NOx concentration before the GPT ozone is turned on, and [NOx]REM is the  
NOx remaining after GPT.1  
2.  
3.  
4.  
Plot the [NO2]CONV concentration output by the instrument on the y-axis against the  
concentration generated [NO2]OUT on the x-axis. The plot should be a straight line within the  
± 2% linearity criteria given for the NOx and NO channels.  
Determine the best straight line fit of the plot either by inspection or least squares. The  
slope of the resulting straight line is the moly efficiency. The value should be between .96  
and 1.02. If not, the moly needs to be replaced.  
If you want the M200AU to automatically compensate (Section 7.6.6.1) for converter  
efficiency, enter the efficiency value in the front panel by CAL-CONC-MOLY-SET, then  
key in the slope from step 3 followed by ENTR. Press EXIT to return to the SAMPLE menu.  
7.6.6.1 Automatic Moly Converter Efficiency Compensation  
The M200AU can automatically compensate the NOx and NO2 readings for the molybdenum  
converter efficiency. There are 2 ways to enter the converter efficiency into the instrument. The  
first is to enter the efficiency as a decimal fraction using the CAL-CONC-MOLY-SET menu.  
The second method is to have the M200AU compute the efficiency using the CAL-MOLY-CAL  
menu. The procedure is given in Table 7-14. To disable the compensation, press CAL-CONC-  
MOLY-SET and enter 1.0000 as the efficiency.  
7-23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 7-14: Automatic Calculation of Converter Efficiency  
Step Number Action  
Comment  
1.  
2.  
3.  
Press  
CAL-CONC-  
CONV-NO2  
From the SAMPLE mode the M200AU enters the Converter  
Efficiency menu and requests expected NO2 cal gas  
concentration.  
Key in expected  
Enter the expected NO2 cal gas concentration. (We suggest you  
NO2 concentration. generate a NO2 concentration of 80% of current range.)  
Then ENTR  
Press  
Reset the previous converter efficiency value to 1.00.  
CAL-CONC-  
MOLY-SET  
Set CE to 1.0000  
4.  
Press  
CAL-CONC-  
CONV-CAL, wait  
Allow the NO2 concentration reading to stabilize. It may take  
time for the ENTR button to appear because the only valid  
values for CE are .96 to 1.02. The ENTR button will not come  
10 min. Then press on unless the value is in this range.  
ENTR  
When ENTR is pressed, the M200AU will compute the ratio of  
observed NO2 concentration to expected NO2 concentration and  
store the ratio.  
5.  
6.  
Press SET  
After the instrument calculates the CE it is good practice to  
check the ratio.  
Press EXIT  
The M200AU will now compensate all readings for this  
converter efficiency value.  
7-24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
7.6.7 Calibration Frequency  
To ensure accurate measurements of the NO, NOx, and NO2 concentrations, calibrate the  
analyzer at the time of installation, and re-calibrate it:  
1. No later than three months after the most recent calibration or performance audit which  
indicated analyzer calibration to be acceptable.  
2. An interruption of more than a few days in analyzer operation.  
3. Any repairs which might affect its calibration.  
4. Physical relocation of the analyzer.  
5. Any other indication (including excessive zero or span drift) of possible significant  
inaccuracy of the analyzer.  
Following any of the activities listed above, the zero and span should be checked to determine if  
a calibration is necessary. If the analyzer zero and span drifts exceed the calibration limits in  
Table 9-1 of Section 2.0.9, Subsection 9.1.3 (Q.A. Handbook), a calibration should be  
performed.  
7.6.8 Other Quality Assurance Procedures  
Precision is determined by a one-point check at least once every two weeks. Accuracy is  
determined by a three-point audit once each quarter.  
Essential to quality assurance are scheduled checks for verifying the operational status of the  
monitoring system. The operator should visit the site at least once each week. Every two weeks a  
Level 1 zero and span check must be made on the analyzer. Level 2 zero and span checks should  
be conducted at a frequency desired by the user. Definitions of these terms are given in  
Table 7-16.  
In addition, an independent precision check between 0.08 and 0.10 ppm must be carried out at  
least once every two weeks. Table 7-16 summarizes the quality assurance activities for routine  
operations. A discussion of each activity appears in the following sections.  
To provide for documentation and accountability of activities, a checklist should be compiled  
and then filled out by the field operator as each activity is completed.  
For information on shelter and sample inlet system, an in-depth study is in Field Operations  
Guide for Automatic Air Monitoring Equipment, Publication No. APTD-0736, PB 202-249 and  
PB 204-650, U.S. Environmental Protection Agency, Office of Air Programs, October 1972.  
7-25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 7-15: Definition of Level 1 and Level 2 Zero and Span Checks  
(from Section 2.0.9 of Q.A. Handbook for Air Pollution Measurement Systems)  
LEVEL 1 ZERO AND SPAN CALIBRATION  
LEVEL 2 ZERO AND SPAN CHECK  
A Level 1 zero and span calibration is a A Level 2 zero and span check is an "unofficial"  
simplified, two-point analyzer calibration used check of an analyzer's response. It may include  
when analyzer linearity does not need to be dynamic checks made with un-certified test  
checked or verified. (Sometimes when no concentrations, artificial stimulation of the  
adjustments are made to the analyzer, the Level 1 analyzer's detector, electronic or other types of  
calibration may be called a zero/span check, in checks of a portion of the analyzer, etc.  
which case it must not be confused with a Level  
Level 2 zero and span checks are not to be used  
2 zero/span check.) Since most analyzers have a  
as a basis for analyzer zero or span adjustments,  
reliably linear or near-linear output response with  
calibration updates, or adjustment of ambient  
concentration, they can be adequately calibrated  
data. They are intended as quick, convenient  
with only two concentration standards (two-point  
checks to be used between zero and span  
concentration). Furthermore, one of the standards  
calibrations to check for possible analyzer  
may be zero concentration, which is relatively  
malfunction or calibration drift. Whenever a  
easily obtained and need not be certified. Hence,  
Level 2 zero or span check indicates a possible  
only one certified concentration standard is  
calibration problem, a Level 1 zero and span (or  
needed for the two-point (Level 1) zero and span  
multipoint) calibration should be carried out  
calibration. Although lacking the advantages of  
before any corrective action is taken.  
the multipoint calibration, the two-point zero and  
If a Level 2 zero and span check is to be used in  
the quality control program, "reference  
span calibration--because of its simplicity--can  
be (and should be) carried out much more  
frequently. Also, two-point calibrations are easily  
automated. Frequency checks or updating of the  
calibration relationship with a two-point zero and  
span calibration improves the quality of the  
monitoring data by helping to keep the  
calibration relationship more closely matched to  
any changes (drifts) in the analyzer response.  
a
response" for the check should be obtained  
immediately following a zero and span (or  
multipoint) calibration while the analyzer's  
calibration is accurately known. Subsequent  
Level 2 check responses should then be  
compared to the most recent reference response  
to determine if a change in response has  
occurred. For automatic Level 2 zero and span  
checks, the first scheduled check following the  
calibration should be used for the reference  
response. It should be kept in mind that any  
Level 2 check that involves only part of the  
analyzer's system cannot provide information  
about the portions of the system not checked and  
therefore cannot be used as a verification of the  
overall analyzer calibration.  
7-26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
7.6.9 Summary of Quality Assurance Checks  
The following items should be checked on a regularly scheduled basis to assure high quality data  
from the M200AU. See Table 7-16 for a summary of activities; also the QA Handbook should be  
checked for specific procedures.  
7-27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 7-16: Activity Matrix for Data Quality  
Frequency And Method  
Of Measurement  
Action If Requirements  
Are Not Met  
Characteristic  
Acceptance Limits  
Shelter temperature  
Mean temperature  
between 22oC and  
28oC (72o and 82oF),  
daily fluctuations not  
greater than ±2oC  
Check thermograph chart  
weekly for variations  
greater than ±2oC (4oF)  
1. Mark strip chart for  
the affected time  
period  
2. Repair or adjust  
temperature control  
Sample introduction  
system  
1. No moisture,  
foreign material,  
leaks, obstructions  
Weekly visual inspection  
Weekly visual inspection  
Clean, repair, or replace  
as needed  
2. Sample line  
connected to  
manifold  
Recorder  
1. Adequate ink &  
paper  
1. Replenish ink and  
paper supply  
2. Legible ink traces  
2. Adjust time to agree  
with clock; note on  
chart  
3. Correct chart  
speed and range  
4. Correct time  
Analyzer operational 1. TEST  
settings measurements at  
Weekly visual inspection  
Adjust or repair as  
needed  
nominal values  
2. M200AU in  
SAMPLE mode  
Analyzer operational Zero and span within Level 1 zero/span every 2  
1. Find source of error  
and repair  
check  
tolerance limits as  
described in Subsec.  
9.1.3 of Sec. 2.0.9  
(Q.A. Handbook)  
weeks  
Level 2 between Level 1  
checks at frequency desired  
by user  
2. After corrective  
action, re-calibrate  
analyzer  
Precision check  
Assess precision as  
described in Sec. 2.0.8 3.4.3 (Ibid.)  
and Subsec. 3.4.3  
Every 2 weeks, Subsec.  
Calc, report precision,  
Sec. 2.0.8 (Ibid.)  
(Ibid.)  
7-28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
7.6.10 ZERO and SPAN Checks  
A system of Level 1 and Level 2 zero/span checks (see Table 7-1) is recommended. These  
checks must be conducted in accordance with the specific guidance given in Subsection 9.1 of  
Section 2.0.9 (Q.A. Handbook). Level 1 zero and span checks must be conducted every two  
weeks. Level 2 checks should be conducted in between the Level 1 checks at a frequency desired  
by the user. Span concentrations for both levels should be between 70 and 90% of the  
measurement range.  
Zero and span data are to be used to:  
1.  
2.  
3.  
Provide data to allow analyzer adjustment for zero and span drift;  
Provide a decision point on when to calibrate the analyzer;  
Provide a decision point on invalidation of monitoring data.  
Items 1 and 2 are described in detail in Subsection 9.1.3 of Section 2.0.9 (Q.A. Handbook). Item  
3 is described in Subsection 9.1.4 of the same section.  
Refer to the Troubleshooting Section 9 of this manual if the instrument is not within the allowed  
variations.  
7.6.10.1 Zero/Span Check Procedures  
The Zero and Span calibration can be checked a variety of different ways. They include:  
1.  
2.  
3.  
4.  
Manual Zero/Span Check  
Zero and Span can be checked from the front panel keyboard. The procedure is in Section 7.1  
of this manual.  
Automatic Zero/Span Checks  
After the appropriate setup, Z/S checks can be performed automatically every night. See  
Table 6-2 and Section 7.3 of this manual for setup and operation procedures.  
Zero/Span checks via remote contact closure  
Zero/Span checks can be initiated via remote contact closures on the rear panel. See Section  
7.5 of this manual.  
Zero/Span via RS-232 port  
Z/S checks can be controlled via the RS-232 port. See Section 5.5 of this manual for more  
details.  
7-29  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
7.6.10.2 Precision Check  
A periodic check is used to assess the data for precision. A one-point precision check must be  
carried out at least once every 2 weeks on each analyzer at an NO2 concentration between 0.08  
and 0.10 ppm. The analyzer must be operated in its normal sampling mode, and the precision test  
gas must pass through all filters, scrubbers, conditioners, and other components used during  
normal ambient sampling. The standards from which precision check test concentrations are  
obtained must be traceable to NIST-SRM. Those standards used for calibration or auditing may  
be used.  
7.6.10.3 Precision Check Procedure  
1.  
Connect the analyzer to a precision gas that has an NO2 concentration between 0.08 and  
0.10 ppm. NO2 precision gas may be generated by either GPT or a NO2 permeation tube. If a  
precision check is made in conjunction with a zero/span check, it must be made prior to any  
zero or span adjustments.  
2.  
3.  
Allow the analyzer to sample the precision gas until a stable trace is obtained.  
Record this value. NO and NOx precision checks should also be made if those data are  
being reported. Information from the check procedure is used to assess the precision of the  
monitoring data; see Section 2.0.8 (Q.A. Handbook) for procedures for calculating and  
reporting precision.  
7.6.11 Recommended Standards for Establishing Traceability  
To assure data of desired quality, two considerations are essential: (1) the measurement process must  
be in statistical control at the time of the measurement and (2) the systematic errors, when combined  
with the random variation in the measurement process, must result in a suitably small uncertainty.  
Evidence of good quality data includes documentation of the quality control checks and the  
independent audits of the measurement process by recording data on specific forms or on a quality  
control chart and by using materials, instruments, and measurement procedures that can be traced to  
appropriate standards of reference. To establish traceability, data must be obtained routinely by  
repeated measurements of standard reference samples (primary, secondary, and/or working  
standards). More specifically, working calibration standards must be traceable to standards of higher  
accuracy, such as those listed below in Table 7-17.  
7-30  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 7-17: NIST-SRM's Available for Traceability of Calibration and Audit Gas  
Standards  
Nominal  
Concentration  
NIST-SRM4  
Type  
Size, at STP  
1683a  
1684a  
1685a  
Nitric oxide in N2  
Nitric oxide in N2  
Nitric oxide in N2  
870  
870  
870  
50 ppm  
100 ppm  
250 ppm  
Permeation Tubes  
Concentration, ppm  
at flow rates of:  
1 L/min/5 L/min  
Permeation rate,  
ug/min  
NIST-SRM4  
Type  
1629  
Nitrogen dioxide  
1.0  
0.5/0.1  
Cylinders of working gas traceable to NIST-SRM's (called EPA Protocol Calibration Gas) are  
also commercially available (from sources such as Scott Specialty Gases, Scott-Marin etc.).  
7.6.12 Certification Procedures of Working Standards  
The NO content of the NO working standard must be periodically assayed against NIST-traceable  
NO or NO2 standards. Any NO2 impurity in the cylinder must also be assayed. Certification of the  
NO working standard should be made on a quarterly basis or more frequently, as required.  
Procedures are outlined below for certification against NO traceable standard. The simplest and most  
straightforward procedure is to certify against a NO standard.  
NOTE  
If the assayed concentration of NO2 impurity in the NO cylinder,  
[NO2]imp, is greater than the 1 ppm value allowed in the calibration  
procedure, make certain that the NO delivery system is not the  
source of contamination before discarding the NO standard.  
7-31  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
7.6.12.1 NO Working Standard Traced to NIST NO Standard  
First, use the NIST-traceable NO standard and the GPT calibration procedure to calibrate the  
NO, NOx, and NO2 responses of the analyzer. Also determine the efficiency of the Molycon.  
Refer to the calibration procedure described in Section 7.6.4.2.  
Then, generate several NO concentrations by diluting the NO working standard. Use the nominal  
NO cylinder concentration, [NO]NOM, to calculate the diluted concentrations. Plot the analyzer  
NO response (in ppm) versus the nominal diluted NO concentration and determine the slope,  
SNOM. Calculate the NO concentration of the working standard [NO]STD from:  
[NO]STD = [NO]NOM x SNOM  
A procedure is presented in the TAD in Reference 2.  
7.6.12.2 Other Methods of Establishing Traceability  
They are:  
1.  
2.  
3.  
NO working standard traced to NIST NO2 standard  
NO2 working standard traced to NIST NO2 standard  
NO2 working standard traced to NIST NO2 standard  
NOTE  
For further information on calibration by GPT and NO2 permeation  
devices, refer to part 50 of Chapter 1, Title 40 CFR, Appendix F  
(revised December 1, 1976) and Reference 13 of that Appendix.  
7.7 Calibration of Independent Ranges or Autoranging  
There are additional considerations when AutoRange or Independent Ranges are selected. The  
M200AU uses one physical range of 0-2000 ppb to cover all EPA concentration ranges. The  
dynamic range of the internal hardware and computer software is sufficient to cover this entire  
range. Internally the range only scales parts of the 0-2000 ppb physical range to cover the  
voltage range selected for the analog outputs.  
7.7.1 Zero Calibration with AutoRange or Independent Range  
Having one physical range to cover all EPA concentration ranges simplifies zero calibration. No  
matter what AutoRange or Independent Range values are selected the values computed from a  
zero calibration are the same. Therefore no special precautions need to be taken when doing a  
zero calibration.  
7-32  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
7.7.2 Span Calibration with AutoRange or Independent Range  
Observe the following guidelines when calibrating in AutoRange or Independent Range.  
1.  
2.  
When doing a span calibration of the M200AU use 80% of the highest of the ranges. This  
will result in the most accurate calibration.  
When selecting concentrations for the NO/NOx or NO2 (GPT) dynamic calibrations use  
80% of the highest AutoRange or the highest Independent Range. This will produce the most  
accurate calibration.  
3.  
4.  
5.  
If the calibration data is obtained from the RS-232 port or from the front panel display,  
no special changes are necessary if the instrument is in AutoRange or Indep. Range. This is  
because the internal hardware and software has sufficient dynamic range to cover the entire  
EPA equivalent range, also those features only affect the analog outputs.  
If using the analog outputs and a chart recorder or datalogger, be sure to note when  
AutoRange occurs so that the correct concentration values are used. With Independent  
Ranges, change the ranges so that all ranges are equal to the highest range so that all  
calibration data is on scale.  
Calibration curves or relationships should be obtained for each range used.  
7.8 Calibration Quality  
After calibration is complete, it is very important to check the QUALITY of the calibration. The  
calibration of the M200AU involves balancing several sections of electronics and software to  
achieve an optimum balance of accuracy, noise, linearity and dynamic range.  
The following procedure checks the Slope and Offset parameters in the equations used to  
compute the NO, and NOx concentrations. It is important that they fall within certain limits with  
respect to themselves and to each other. For an explanation of the use of these terms in the  
concentration calculation see Section 5.2.2.5.  
The slope and offset parameters are similar to the span and zero pots on an analog instrument.  
Just as in the analog instrument, if the slope or offset get outside of a certain range, the  
instrument will not perform as well.  
The slope value will be slightly different on the NO and NOx channels. This is due to slight  
differences in pneumatic resistance in each pathway. If the slopes are significantly different,  
there is a calibration error or a cross port leak in the switching valve. If there is a sudden large  
change in slopes after a calibration, that usually indicates a change in reaction cell pressure. This  
generally requires a Factory Calibration covered in Section 9.1.6.  
7-33  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
The offset value gives information about the background signal level. Check the observed offset  
value against the factory value in Table 2-1. If significantly higher, check Section 9.1.6. Also,  
after calibration check the PREREACT – TEST function reading and compare it against the  
factory checkout value. Increasing readings are a predictor of problems.  
Table 7-18: Calibration Quality Check  
Step Number Action  
Comment  
1.  
Scroll the TEST  
The SLOPE value for NOx should be 1.0 ± 0.3. If the value is  
not in this range, check Section 9.1.6. If the SLOPE value is in  
the acceptable range the instrument will perform optimally.  
function menu until  
the NOx SLOPE is  
displayed.  
2.  
Scroll the TEST  
function menu until  
the NO SLOPE is  
displayed.  
The SLOPE value for NO should be 1.0 ± 0.3. If the SLOPE is  
in the acceptable range the instrument will perform optimally. If  
the value is not in this range, check Section 9.1.6.  
NOTE:  
The NO and NOx slopes should be equal within ± 0.1.  
3.  
4.  
Scroll the TEST  
This number should be near zero. Values between –10 to 150  
function menu until mV are acceptable. This number already has the PREREACT  
the NOx OFFSET is value subtracted out and is mainly the background signal due to  
displayed.  
the molybdenum converter. If the OFFSET value is outside this  
range, check Section 9.1.6.  
Scroll the TEST  
The instrument will now display the NO OFFSET value. Values  
function menu until between –10 to 150 mV are acceptable. This number already  
the NO OFFSET is has the PREREACT reading subtracted out and should be near  
displayed.  
zero. If the OFFSET value is outside this range, check  
Section 9.1.6.  
After the above procedure is complete, the M200AU is ready to measure sample gas.  
7-34  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
7.9 References  
1.  
Environmental Protection Agency, Title 40, Code of Federal Regulations, Part 50,  
Appendix F, Measurement Principle and Calibration Procedure for the Measurement of  
Nitrogen Dioxide in the Atmosphere (Gas Phase Chemiluminescence), Federal Register, 41  
(232): pp 52688-52692, December 1976.  
2.  
Ellis, Elizabeth C. Technical Assistance Document for the Chemiluminescence  
Measurement of Nitrogen Dioxide, U.S. Environmental Protection Agency, Research  
Triangle Park, NC. October 1976. 91 p.  
3.  
4.  
5.  
6.  
7.  
Quality Assurance Requirements for State and Local Air Monitoring Stations (SLAMS),  
Appendix A, Federal Register, Vol. 44, No. 92, pp 27574-27582, May 1979.  
Catalog of NBS Standard Reference Materials. NBS Special Publication 260, 1975-76  
Edition. U.S. Department of Commerce, NBS. Washington, D.C. June 1975.  
Quality Assurance Handbook for Air Pollution Measurement Systems - Volume I,  
Principles. EPAN-600/9-76-005. December 1984.  
Quality Assurance Handbook for Air Pollution Measurement Systems - Volume II,  
Ambient Air Specific Methods. EPA-600/4-77/027a, December 1986.  
Quality Assurance Requirements for Prevention of Significant Deterioration (PSD) Air  
Monitoring, Appendix B, Federal Register, Vol. 44, No. 92, pp 27582-27584, May 1979  
7-35  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
INTENTIONALLY BLANK  
7-36  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
8 MAINTENANCE  
8.1 Maintenance Schedule  
NOTE  
The operations outlined in this chapter are to be  
performed by qualified maintenance personnel only.  
Table 8-1: Preventative Maintenance Schedule  
Item  
Maintenance Interval  
6 - 12 month intervals  
Annually or after repairs  
Daily  
Reference Section  
Table 9-1  
TEST Functions  
Zero/Span Calibration  
Zero/Span Checks  
Particulate Filter  
Filter for PermaPure Drier  
Reaction Cell Window  
Ozone Flow  
Section 7  
Section 7, Table 6-2  
Figure 8-1  
Weekly as needed  
Replace every 12 months  
Clean annually or as necessary  
Check every year  
Figure 9-7  
Section 9.3.8, 8.4, Figure 8-3  
Figure 9-9, Section 9.3.7  
Figure 9-9, Section 9.3.7  
Figure 8-4, Section 8.5  
Sample Flow  
Check every year  
Moly Converter  
Check efficiency every 6  
months  
Pneumatic Lines  
Examine every 12 months, clean Figure 8-5, Figure 8-6, Figure 8-7  
if necessary  
Factory Calibration  
Calibrate each year or after  
repairs  
Section 9.1.6  
Leak Check  
O-rings  
Check every year  
Section 8.7  
Replace if seal is broken  
Figure 9-9, Figure 9-10  
8-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 8-2: Preventative Maintenance Calendar  
Maintenance  
Interval  
Item  
Jan Feb Mar Apr May Jun Jul  
Aug Sep Oct Nov Dec  
6 - 12 month  
intervals  
TEST Functions  
Zero/Span  
Calibration  
Annually or after  
repairs  
Zero/Span Checks  
Particulate Filter  
Daily  
Weekly as needed  
Filter for PermaPure  
Drier  
Replace every 12  
months  
Reaction Cell  
Window  
Clean annually or as  
necessary  
Ozone Flow  
Sample Flow  
Check every year  
Check every year  
Check efficiency  
every 6 months  
Moly Converter  
Examine every 12  
months, clean if  
necessary  
Pneumatic Lines  
Calibrate each year  
or after repairs  
Factory Calibration  
Leak Check  
O-rings  
Check every year  
Replace if seal is  
broken  
8-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
8.2 Replacing the Sample Particulate Filter  
The particulate filter should be inspected often for signs of plugging or contamination. It is also  
common for dirt particles to absorb NO2, thus causing those readings to be low.  
To check and change the filter:  
1.  
2.  
Fold down the M200AU front panel.  
Locate the filter on the left side of the analyzer front panel. See Figure 8-1 for an  
exploded view of the filter assembly.  
3.  
4.  
Visually inspect the filter through the glass window.  
If the filter appears dirty, unscrew the hold-down ring, remove the teflon o-ring and then  
the filter.  
5.  
Replace the filter, being careful that the element is fully seated in the bottom of the  
holder. Replace the teflon o-ring, then screw on the hold-down ring and hand tighten.  
8-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Figure 8-1: Replacing the Particulate Filter  
8-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
8.3 Sample Pump Maintenance  
The external sample pump is capable of maintaining the cell pressure at better than 4.0 "Hg-A. If  
a higher pressure is noted, the pump may need servicing. Check the pump and pneumatic system  
for leaks or rebuild pump. The ozone scrubber is integrated into the molybdenum converter  
inside the instrument.  
See Figure 8-2 for component locations.  
8-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Figure 8-2: Sample Pump Assembly  
8-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
8.4 Cleaning the Reaction Cell  
The reaction cell should be cleaned whenever troubleshooting points to it as the cause of the  
trouble. A dirty cell will cause excessive noise, unstable zero or span, or low response.  
To clean the reaction cell it is necessary to remove the reaction cell from the sensor housing. Use  
the following guide:  
1.  
2.  
Turn off the instrument power and vacuum pump.  
Loosen the hold down screws for pneumatic sensor assembly and move the assembly to  
the side.  
3.  
4.  
5.  
6.  
Disconnect the exhaust fitting and inlet fittings. See Figure 8-3.  
Loosen four screws holding the reaction cell to the sensor.  
Disconnect heater/thermistor and lift the cell away.  
The reaction cell will separate into two halves:  
A.  
B.  
The manifold assembly  
The reaction block with reaction sleeve and window  
7.  
8.  
The reaction sleeve and window should be cleaned with methanol and a clean tissue and  
dried.  
Normally it is not necessary to clean the sample and ozone flow orifices since they are  
protected by sintered SS filters. If tests shows that cleaning is necessary then do the  
following:  
A.  
The manifold O-rings, springs, sintered SS filters, and orifices should be removed  
before cleaning. It is suggested that the orifice, filter and o-rings be replaced  
unless an ultrasonic cleaner and methanol or methylene chloride is available. Both  
orifice and sintered filter may be cleaned with an ultrasonic bath for 30 minutes in  
either solvent.  
B.  
C.  
D.  
After cleaning with solvent, flush copiously with tap waster.  
Now, rinse parts in either D1 or distilled water.  
Dry parts prior to re-installation.  
9.  
Do not remove the sample and ozone nozzles. They are Teflon threaded and require a  
special tool for re-assembly. If necessary, the manifold with nozzles attached can be  
cleaned in an ultrasonic bath.  
10.  
11.  
Reassemble in proper order and re-attach onto sensor housing. Reconnect pneumatics and  
heater connections, then re-attach the pneumatic sensor assembly and the cleaning  
procedure is complete.  
After cleaning, the analyzer span response may drop 10 - 12% in the first 1-2 days as the  
reaction cell window conditions.  
8-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Figure 8-3: Reaction Cell Assembly  
8-8  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
8.5 Replacing the Molybdenum Converter  
The Molybdenum Converter is located in the center of the instrument, refer to Figure 2-5 for its  
location. Refer to Figure 8-4. The heater, thermocouple, converter is designed to be replaced as a  
single unit.  
1.  
2.  
Turn off the power to the M200AU and allow the converter to cool.  
Remove the entire assembly from the chassis  
A. Remove the pneumatic fittings from the valves.  
B. Remove the power to the valves, thermocouple wire and cartridge heater.  
C. Remove the converter assembly from the chassis by loosening the 4 captive screws that  
secures the assembly to the chassis.  
CAUTION  
The converter operates at 315ºC. Severe burns can result  
if not enough time is allowed for the assembly to cool. Do  
not handle assembly until it is at room temperature.  
3.  
4.  
Disconnect the gas fittings and power cable grounding from the can.  
Remove the valve assembly and bottom bracket and re-attach those two parts to the  
replacement moly assembly.  
5.  
6.  
Re-attach the pneumatic fittings and valve assembly to the can.  
Install the assembly back into the analyzer. Re-attach the electrical and pneumatic  
fittings. Leak check the assembly when completed.  
7.  
Turn the power back on. The insulation can emit a burnt odor for the first 24 hours, this is  
normal. Allow the converter to burn-in for 24 hours, then re-calibrate the instrument.  
8-9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Figure 8-4: Molybdenum Converter Assembly  
8-10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
8.6 Pneumatic Line Inspection  
Particulate matter in the pneumatic lines will affect both flow rate and response time. It is  
important that the pneumatic system be periodically inspected and thoroughly cleaned if  
necessary. Clean by disassembling and passing methanol through three times. Dry with nitrogen  
or clean zero air.  
Also inspect all pneumatic lines for cracks and abrasion on a regular basis. Replace as necessary.  
Refer to the pneumatic diagram in Figure 8-5, Figure 8-6 and Figure 8-7.  
8-11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Figure 8-5: Pneumatic Diagram - Standard Configuration  
8-12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Figure 8-6: Pneumatic Diagram with Zero/Span Valve Option  
8-13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Figure 8-7: Pneumatic Diagram with External Converter Option  
8-14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
8.7 Leak Check Procedure  
If a leak checker is not available, it is possible to leak check the instrument using the M200AU's  
pump plus a shut-off valve.  
1.  
2.  
Turn off instrument power and pump power.  
Cap the sample inlet port, ozone generator air inlet, and zero air inlet (if Z/S valve option  
present).  
3.  
4.  
Insert a shut-off valve between the sample pump and the vacuum manifold at the rear of  
the instrument.  
Turn on the sample pump and set the TEST function to RCEL, which measures the  
reaction cell pressure. Close the shutoff valve and monitor the cell pressure. The pressure  
should not drop more than 1"-Hg (0.5psi) in 5 minutes. If there is a leak, it is not possible by  
this method to tell where it is located. You can locate the leak by using a pressure leak  
checker described below.  
5.  
The sensor module is equipped with a fitting at the top of housing near the heat sink fins.  
This fitting can be pressurized and the sensor checked for leaks. The leak-down rate is the  
same as above.  
If you have a leak checker:  
1.  
2.  
Turn off instrument power and pump power.  
Disconnect pump at rear panel. Cap the sample inlet port, ozone generator air inlet, and  
zero air inlet (if Z/S valve option present) and connect the leak checker to the exhaust port.  
CAUTION  
Pressure must be less than 15 psi.  
8-15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
3.  
Pressurize system and check for leaks by watching overall pressure. The pressure should  
not drop more than 1"-Hg (0.5psi) in 5 minutes.  
If the instrument fails the pressure test, each fitting needs to be leak checked to find the location.  
Be careful that the system is always pressurized so as not to draw soap solution into the  
plumbing system. Make sure you dry off any accumulated bubble solution. Refer to Figure 8-5,  
Figure 8-6 and Figure 8-7 for pneumatic diagrams.  
The Sensor module can be leak checked as a unit using a 1/8" tubing fitting on top of the  
assembly. The same rules as above apply.  
1.  
2.  
Pressurize to <15psi.  
After turning off pressure tester, pressure should not drop more than 1" Hg in 5 minutes.  
8.8 Light Leak Check Procedure  
1.  
2.  
3.  
Scroll the TEST functions to PMT.  
Input zero gas  
Shine a powerful flashlight or portable incandescent light at the inlet and outlet fitting,  
and at all the joints of the reaction cell. The PMT value should not respond to the light.  
If there is a response, tighten the joints or replace the tubing with new black PTFE tubing.  
We often find light leaks are caused by o-rings being left out of the assembly.  
8.9 Prom Replacement Procedure  
Preparation: If any setup changes such as RANGE, AUTOCAL ON/OFF etc. have been made,  
record the changes because all settings should be checked after the PROM is changed. See  
Figure 9-2 for location of prom on CPU card.  
1.  
2.  
Turn the machine off.  
Remove the hold down screw that holds in the V/F-CPU assembly to the motherboard.  
Disconnect the J9 power connector from the motherboard. Gently lift the assembly far  
enough out of the instrument to remove the connector to the display and the RS-232  
connector.  
3.  
The CPU board is attached to the larger V/F board.  
8-16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
4.  
Remove the board, laying it down on an insulating surface such that the board edge pins  
on the PCB are on the left. The PROM chip should be at the top center. The current chip  
should be labeled with something like "2AUC8…". See Figure 9-2 for prom location. Gently  
pry the chip from its socket and replace it with the new chip. Install the chip in the left end of  
the socket with the notch facing to the right. Make sure that all of the legs insert into the  
socket correctly.  
5.  
6.  
Re-attach the CPU board to the V/F board, and re-attach the assembly to the  
motherboard.  
Turn the M200AU ON and observe the front panel display. As the machine goes through  
the setup the version number will be displayed on the front panel. It should read the same as  
the version number printed on the prom.  
7.  
8.  
Re-enter any non-default settings such as RANGE or AUTOCAL. Re-enter the SPAN  
values in the CAL-CONC menu. Check all settings to make sure that expected setup  
parameters are present.  
Re-calibrate the Analyzer so that the default slope and intercept are overwritten with the  
correct values.  
8-17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
INTENTIONALLY BLANK  
8-18  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
9 TROUBLESHOOTING, ADJUSTMENTS  
NOTE  
The operations outlined in this chapter are to be  
performed by qualified maintenance personnel only.  
This section of the manual contains information on diagnosing and repairing instrument  
performance problems. It contains information on how to use and interpret TEST and  
DIAGNOSTIC data as well as WARNING messages the instrument generates. There is  
information on how to troubleshoot the instrument subsystems. Finally there is information to  
perform adjustments such as DAC calibration procedures.  
This manual provides troubleshooting procedures that address problems to the board level. For  
component level troubleshooting, consult the schematics for the appropriate board in the  
Appendix.  
NOTE  
The values of the readings shown on the front panel of the  
instrument may at times read XXXXXX. This means that the  
reading is off scale and therefore meaningless.  
General Troubleshooting Hints  
1.  
2.  
If the fault light is on and it stays on after you clear the warning messages, see Section  
9.1.2.  
Think of the analyzer as three sections:  
Section 1: Pneumatics - Over 50% of all analyzer problems are traced to leaks in the pump  
assembly, sample filter, instrument internal pneumatics, calibrator or external sample  
handling equipment. Suspect a leak first, and refer to Section 8.7.  
Section 2: Electronics - data processing section. This can be readily checked out using  
Electric Test in Section 9.1.3.2.  
Section 3: Optics - Optical section consisting of PMT, HVPS, Preamp, and signal  
processing. Refer to Section 9.1.3.3 on use of Optical Test.  
9-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
3.  
Check the TEST functions:  
A. Compare the TEST functions to the factory values in Table 2-1. This will often provide  
important clues as to the problem.  
B. Pay special attention to the NO and NOx slopes:  
C. The slopes are the software equivalent of the span pot on an analog instrument. If the  
slopes are not 1.0 ± 0.3, the gain has changed.  
D. Check the Pre-reactor - PREREACT - reading in the TEST functions. Compare it to the  
value in the factory checkout Table 2-1. If the reading is significantly greater than the  
factory test value, the reaction cell could be contaminated or there could be a light leak in  
the cell. Verify this fault by turning the ozone generator off and see if the reading drops  
more than 25 mV.  
E. Check for a change in cell pressure (vacuum) - compare to value in Table 2-1. - possible  
causes:  
1)  
2)  
3)  
4)  
Partially plugged ozone killer  
Change of pump or malfunctioning pump  
Plugged pneumatics  
Change in altitude  
4.  
Check for pneumatic leaks - perform the leak check procedure in Section 8.7. If slopes  
are different from each other by > .1, this usually indicates a leak in the switching (NO/NOx)  
valve or improper calibration.  
5.  
6.  
7.  
Check for light leaks - Turn off the ozone generator, then wait 7 minutes. If the reading  
drops significantly, the reaction cell is contaminated. If not, a light leak is indicated.  
Incorrect span gas concentration - this could come either from the calibrator or entering  
the expected span gas concentration in the M200AU incorrectly, see Table 7-3.  
If the instrument does not respond to span gas, check Section 9.2.3.  
The above should get you started in diagnosing and repairing the most common faults. If these  
reasons have been eliminated, the next thing to do is a Factory Calibration covered in Section  
9.1.6 or check Section 9.2 for other fault diagnosis. If difficulties persist, contact our service  
department. The 800 telephone number is on the cover page of this manual.  
9-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
9.1 Operation Verification-M200AU Diagnostic Techniques  
9.1.1 Fault Diagnosis with TEST Variables  
Table 9-1 indicates possible fault conditions that could cause the TEST functions to be outside  
the acceptable range.  
9-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 9-1: Test Functions  
Test Function  
Factory Set-Up  
Comment  
RANGE  
500 PPB  
This is the Range of the instrument. In standard configuration all 3  
outputs have the same range.  
Independent range option allows different ranges for each output.  
When enabled, there will be 3 range values displayed.  
Auto range option allows 2 different ranges for all outputs, and  
will automatically switch to the other range dynamically as  
concentration values require. The TEST values will show the  
range the instrument is currently operating in, and will  
dynamically display the alternate range as the range changes  
occur.  
STABIL  
Check value in  
Final Test Values  
Table 2-1  
The instrument stability is the Std. Deviation of the last 50 min of  
NOx conc data. It is computed for the NOx channel only. The noise  
value only becomes meaningful if sampling a constant  
concentration for more than 50 minutes. The noise value should be  
compared to the value observed in the factory check-out.  
Faults that cause high noise values are:  
1. Gas leaks  
2. Light leak  
3. Faulty HVPS  
4. Defective Preamp board  
5. Outgassing Moly converter  
6. PMT recently exposed to room light  
7. Dirty/contaminated reaction cell  
8. Mis-calibrated (slope - offset outside of limits)  
SAMPLE FLW  
1000 cc/min  
± 100  
This is the instrument flow. It is computed using the up stream and  
down stream pressures across the sample flow orifice. This  
method can give a false flow indication if the orifice is plugged  
and the sample pump is creating a pressure drop. It should be  
taken into account when diagnosing instrument faults.  
- A rapid method of determining if the orifice is plugged is to  
disconnect the sample and ozone tubes from the reaction cell, then  
briefly put your finger over the fittings on the cell. You should  
feel the vacuum build up. Also note the difference between the  
high sample flow and the low ozone flow rate.  
- Another reliable method is to attach a rotameter or soap bubble  
flowmeter to the fittings to measure the flows.  
Flow rate will change ± a few cc/min due to changes in ambient  
air pressure such as cycling of air conditioning, or passing weather  
fronts. Changing altitude changes the ambient air pressure and  
therefore the sample flowrate. This effect is about 15-20 cc/min  
per 1000 feet of altitude change. If required, the output of the  
instrument can be compensated for pressure. See Section 5.3.9,  
Table 9-5.  
(table continued)  
9-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 9-1: Test Functions (Continued)  
Test Function  
Factory Set-Up  
Comment  
OZONE FL  
This is the Ozone flow. It is measured by a solid state flow meter,  
and thus is a true indication of flow.  
80 cc/min ± 10  
If you suspect there is no ozone being generated, disconnect the  
tube at the reaction cell and rub the end of the tube on your  
fingertips, then sniff your fingers. The odor of ozone should be  
readily apparent.  
PMT  
0-5000 mV  
This is the instantaneous output of the PMT. During normal  
operation the value varies widely as the M200AU switches from  
NO to NOx to Pre-reactor modes. Changes in reading will be  
synchronized with valve switching. The PMT voltage values will  
be relatively constant when:  
1. Electric test - variation in the 2000 mV signal observed will be  
sampling errors of the V/F board and preamp noise. See Section  
9.1.3.2.  
2. Optic test - variation in the 2000 mV signal will be PMT dark  
current, preamp, HVPS plus item 1 above. See Section 9.1.3.3.  
3. Sampling zero gas - signal from 1, 2 plus signal from ozone  
generator air  
4. Sampling pure NO span gas - signal will be 1, 2, 3, above plus  
signal from chemiluminescent reaction. Slight pulsations will  
be noticed as the M200AU switches from NO to NOx. This is  
due to differences in flowrates in each channel. These  
differences are taken out in the calibration process resulting in  
slightly different slopes for the NO and NOx channels. Large  
pulsations when switching to the NOx channel is indicative of a  
bad moly converter.  
When sampling zero gas the PMT reading should be less than 150  
mV and relatively constant.  
High or noisy readings could be due to:  
1. Excessive background light which is caused by a possible  
contaminated reaction cell.  
2. Humidity (undried ambient air) in the ozone generator feed air.  
3. PMT recently exposed to room light. It takes 24-48 hours for  
the PMT to adapt to dim light.  
4. Light leak in reaction cell.  
5. Improper slopes  
NORM PMT  
0-5000 mV  
The Normalized PMT reading is to be used as the PMT reading  
during the FACTORY CALIBRATION procedure. In addition to  
the raw mV reading from the PMT this reading is adjusted using  
certain other factors to produce an accurate reading for the Factory  
Calibration procedure.  
(table continued)  
9-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 9-1: Test Functions (Continued)  
Test Function  
Factory Set-Up  
Comment  
PREREACT  
Check value in  
Final Test Values  
Table 2-1  
PREREACT is the current value of the Pre-reactor circuit reading.  
The number typically should be  
0 mV -15/+35 if the instrument is sampling zero air. When the  
M200AU is in SAMPLE mode readings may be higher if sample  
gas contains hydrocarbons. If a problem is suspected, check the  
Pre-reactor and NO/NOx valves for cross port leaks. Use  
Diagnostic mode to manually switch the valves while checking for  
leaks and correct operation.  
HVPS  
450 – 900 VDC  
This represents the scaled-up HVPS programming voltage to the  
HVPS. The design of the HVPS precludes taking a single reading  
that indicates the health of the supply. Refer to the HVPS  
Troubleshooting Section 9.3.8.5 for a procedure for testing the  
HVPS. This TEST function is used primarily to set the HVPS  
voltage value. A value not in the 450 to 900 volt range indicates  
problems with the HVPS supply.  
DCPS  
2500 ± 200 mV  
DCPS is a composite of the +5 and ± 15 VDC supplies. It has  
been arbitrarily set at 2500 ± 200 mV. If it is not in this range one  
of the voltages in the supply is not working. Check the procedures  
for diagnosing the Power Supply Module in Section 9.3.5.  
RCELL TEMP  
40 ± 1° C  
The reaction cell temperature is controlled to 40° C ± 1° C by the  
computer. It should only read other values when the instrument is  
warming up. If the value is outside the acceptable range, go to the  
procedure for diagnosing the Reaction cell temp supply in Section  
9.3.8.2. The alarm limits are less than 35° C and greater than  
45° C.  
BOX TEMP  
PMT TEMP  
The Box Temp is read from a thermistor on the Status/Temp  
board. It should usually read about 5° C above room temp. The  
M200AU is designed to operate from 20 to 30° C ambient.  
Therefore the box temperature should be in the range of about 10  
to 50° C. Temperatures outside this range will cause premature  
failures of components, and poor data quality. Warning limits are  
< 8° C and > 48° C.  
8 - 48° C  
The PMT detector is very temperature sensitive. The PMT  
temperature should always be –5° C, except at power-up.  
Temperatures more than ± 1° C from the set point indicate  
problems with the cooler circuit. See Section 9.3.8.4 for PMT  
cooler diagnostic and troubleshooting. Warning limits are < -8° C  
and > -2° C.  
-5 ± 1° C  
40 ± 2o C  
BLOCK TEMP  
Reports the temperature of the sample flow and ozone flow  
control orifices. Variations in these temperatures will cause span  
concentration drift.  
(table continued)  
9-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 9-1: Test Functions (Continued)  
Test Function  
Factory Set-Up  
Comment  
MOLY TEMP  
Moly temp is controlled by the CPU to 315° C. After cold start it  
requires about 30 min to come to temperature. After temperature is  
315 ± 5° C  
reached temperature should not vary more than ± 5° C. See  
Section 9.3.4 for troubleshooting. Warning limits are < 290° C and  
> 320° C.  
RCEL PRESS  
RCEL is the pressure in the Reaction Cell. The instrument is very  
sensitive to variations in reaction cell pressure. A 10% change in  
output per 1"Hg pressure is typical. The pressure reading will  
change when the NO/NOx and Pre-reactor valves switch,  
otherwise it should remain constant. The reaction cell pressure  
will be lower at higher altitude due to lower pump back-pressure.  
Pressures higher than the acceptable range will decrease  
instrument noise performance and sensitivity.  
3.5 ± 1 in-Hg-A  
SAMP PRESS  
29.5"Hg at sea  
level  
The sample pressure is taken upstream of the reaction cell. It  
usually runs about 0.5" less than ambient pressure due to the  
restrictions in the sample intake tubing. Sample pressure should be  
within ± 1"Hg of atmospheric pressure. The pressure sensor used  
reports absolute pressure and therefore is sensitive to altitude,  
weather fronts, and room air conditioning. Changes due to altitude  
is about 1" per 1000 ft., other changes are ± 0.4" maximum.  
Pressurizing the sample inlet will cause the M200AU to be noisy  
and to shift its reading.  
SLOPE  
OFFSET  
TIME  
This is the software slope value. It operates like a software gain  
pot. Refer to Table 7-18 on Calibration Quality for additional  
information.  
1.0 ± 0.3  
0 – 10, +150  
This is the software offset value. It operates like a software DC  
offset pot. Refer to Table 7-18 on Calibration quality for  
additional information.  
HH:MM:SS  
This is the time of day clock readout. It is used to time the  
AutoCal cycles. The speed of the clock can be adjusted by the  
CLOCK_ADJ variable in the VARS menu. The clock can be set  
via SETUP-CLOCK-TIME from the front panel.  
9-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
9.1.2 Fault Diagnosis with WARNING Messages  
The M200AU monitors several internal values for alarm conditions. If the condition for an alarm  
is met, the alarm is displayed on the front panel and the warning is transmitted out the RS-232  
port. If a warning is present, it can be cleared by pressing the CLR key on the front panel. If  
uncleared warnings are present they can be examined by pressing the MSG button on the  
keyboard. Any time the instrument is powered up, the SYSTEM RESET alarm will be displayed.  
Generally, it is ok to ignore warnings that are displayed shortly after power-up, only if they  
persist should they be investigated.  
Table 9-2 shows the warning messages and gives some possible causes.  
9-8  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 9-2: Front Panel Warning Messages  
Message  
Description  
SYSTEM RESET  
Analyzer was reset/powered on. This warning occurs every  
time the instrument is powered up, as in a power failure. It can  
also occur if the RAM or EEPROM is reset.  
RAM INITIALIZED  
RAM was erased. The RAM contains the DAS averages which  
get erased when the RAM is initialized. It also contains  
temporary data used by the M200AU to calculate  
concentrations. No setup variables are stored in the RAM.  
SAMPLE FLOW WARN  
OZONE FLOW WARNING  
The calculated sample flow is outside the hi/low limits. Since  
the flow is calculated, it probably means the pressure has  
gotten too low. Flow warnings are caused by a plugged sample  
inlet or ozone line.  
Ozone flow out of spec. Warnings occur most often because of  
loss of vacuum, which causes the ozone flow to go to zero.  
They also can occur due to a flow sensor failure.  
RxCELL PRESS WARNING  
BOX TEMP WARNING  
Vacuum out of spec. Warnings are caused by leaks, pump  
failure or disconnected pump.  
Box temp. out of spec. Instrument fan failure, enclosure  
temperature failure. Operation of the M200AU in a too warm  
or cold environment will cause degradation of data quality and  
shorten the life of the instrument.  
RCELL TEMP WARNING  
PMT TEMP WARNING  
Reaction cell temp. out of spec. The warning message is most  
often present during initial warm-up or if the connector to the  
heaters is not plugged in after dis-assembly. It has also  
occurred if the thermistor is not in position in the reaction cell.  
PMT temp. out of spec. The PMT temp has its own  
proportional controller on the preamp board. Warnings might  
occur during warmup. A warning can occur if the 7 pin  
connector to the interior of the sensor is not plugged in. The  
power connector to the PSM should be checked for proper  
voltage (+15 VDC ± 0.5)  
MOLY TEMP WARNING  
Molycon temp. out of spec. The Moly temp is controlled by  
the CPU. It has a thermocouple with amplifier on the  
Status/Temp board. Because of the high temperature of the  
Moly (315° C), the moly temp warning will tend to be the last  
warning to clear as the instrument is powered on.  
(table continued)  
9-9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 9-2: Front Panel Warning Messages (Continued)  
Message  
Description  
CANNOT DYN ZERO  
Dynamic zero cal. out of spec. The reading of the PMT was  
too high for the ZERO button to appear. Make sure the  
instrument is receiving zero gas. Check for dirty reaction cell.  
Do the factory calibration procedure located in Section 9.1.6.  
CANNOT DYN SPAN  
Dynamic span cal. out of spec. The reading of the PMT was  
too high or low for the SPAN button to appear. Make sure the  
instrument is receiving correct concentration span gas. Make  
sure the expected span concentration is entered. Check for  
dirty reaction cell. Do the factory calibration procedure located  
in Section 9.1.6.  
OZONE GEN OFF  
Ozone generator is off. See Table 9-11 for conditions.  
PRACT WRN XXX.X MV  
The Pre-reactor circuit compensates for detector dark current,  
and background light. Certain electrical faults cause high  
readings to be added to the filter. First, the cause of the high  
PREREACT reading should be found and repaired, then wait  
10 minutes for the Pre-reactor filter to clear itself out.  
A/D NOT INSTALLED  
V/F board has failed. The V/F board did not respond to  
commands from the CPU. This probably means 1. board not  
seated in socket 2. defective board 3. defective back plane  
connector  
HVPS WARNING  
DCPS WARNING  
High Voltage Power Supply voltage out of limits. Limits are  
400-900 VDC  
DC Power Supply voltage out of limits. Limits are 2500  
± 500 mV  
9-10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
9.1.3 Fault Diagnosis using DIAGNOSTIC Mode  
Diagnostic mode can be looked at as a tool kit of diagnostics to help troubleshoot the instrument.  
To enter DIAG mode press SETUP-MORE-DIAG.  
The diagnostic modes are summarized in Table 9-3. To access these functions, after you have  
pressed SETUP-MORE-DIAG, press NEXT, PREV to select the desired mode, then press  
ENTR.  
9-11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 9-3: Summary of Diagnostic Modes  
DIAG Mode  
Description  
SIGNAL I/O  
Gives access to the digital and analog inputs and outputs on  
the V/F board. The status or value of all of the signals can be  
seen. Some of the signals can be controlled from the keyboard.  
Table 9-4 gives details on each signal and information on  
control capabilities.  
NOTE:  
Some signals can be toggled into states that indicate warnings  
or other faults. These settings will remain in effect until DIAG  
mode is exited, then the M200AU will resume control over the  
signals.  
ANALOG OUTPUT  
D/A CALIBRATION  
Causes a test signal to be written to the analog output DAC's.  
The signal consists of a scrolling 0%, 20%, 40%, 60%, 80%,  
100% of the analog output value. The scrolling may be  
stopped by pressing the key underneath the % display to hold  
the current value. The exact voltage values depend on the  
switch settings on the V/F board amplifiers.  
The analog output is created by 4 digital-to-analog converters.  
This selection starts a procedure to calibrate these outputs.  
Refer to Section 9.3.3.1 for a detailed procedure.  
TEST CHANNEL  
OPTICAL TEST  
Allows several different internal voltages to be routed to an  
analog output port. Useful for diagnosing intermittent faults.  
Sets the M200AU into a known state and turns on an LED  
near the PMT to test the instrument signal path. See Section  
9.1.3.3 for details on using this test.  
ELECTRICAL TEST  
O3 GEN OVERRIDE  
RS-232  
Tests just the electronic portion of the PMT signal path. Used  
in conjunction with optic test, see Section 9.1.3.2.  
This function switches the power to the ozone generator. It  
does not indicate the OFF/ON status of the generator  
Causes a 1 second burst of data to be transmitted from the RS-  
232 port. Used to diagnose RS-232 port problems. See Section  
9.1.3.7, 9.3.2 for RS-232 port diagnostic techniques.  
9-12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
9.1.3.1 Signal I/O Diagnostic  
Table 9-4: Diagnostic Mode - Signal I/O  
No.  
Signal  
Control  
Description  
0
DSP_BROWNOUT  
NO  
Display brownout is used to keep the display from  
getting corrupted during low line voltage conditions.  
Circuitry on the Status/Temp board senses low line  
voltage and sets this bit. The CPU reads this and  
generates the BROWNOUT_RST signal described  
below.  
1
2
EXT_ZERO_CAL  
EXT_SPAN_CAL  
NO  
NO  
Shows state of status input bit to cause the M200AU  
to enter Zero Calibration mode. Use to check external  
contact closure circuitry.  
Shows state of status input bit to cause the M200AU  
to enter the Span Calibration mode. Use to check  
external contact closure circuitry.  
3
4
5
6
SPAN_VALVE  
CAL_VALVE  
YES  
YES  
YES  
YES  
Switches the Zero/Span valve. Use this bit to test the  
valve function.  
Switches the Sample/Cal valve. Use this bit to test  
the valve function.  
NOX_VALVE  
RCELL_HEATER  
Switches the NO/NOx valve. Use this bit to test the  
valve function.  
Shows the status of the reaction cell heater. This has  
the same function as the LED in the Power Supply  
Module.  
7
8
BLOCK_HEATER  
ELEC_TEST  
YES  
YES  
Shows the status of the block heater.  
Turns on electric test bit on the preamp board. Should  
be used for troubleshooting Preamp logic lines. We  
recommend you use the ELEC TEST button in the  
DIAG menu to operate electric test.  
9
OPTIC_TEST  
YES  
YES  
Turns on optic test bit in preamp. Should be used for  
troubleshooting Preamp logic lines.  
We recommend you use the OPTIC TEST button in  
the DIAG menu to operate optic test.  
10  
BROWNOUT_RST  
Brownout reset works in conjunction with  
DSP_BROWNOUT. When DSP_BROWNOUT is  
set, the CPU sends a signal to reset the display and  
clear the DSP_BROWNOUT.  
(table continued)  
9-13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 9-4: Diagnostic Mode - Signal I/O (Continued)  
No.  
Signal  
Control  
Description  
11  
CONV_HEATER  
YES  
Shows the status of the Moly heater. This has the  
same function as the LED in the Power Supply  
Module.  
12  
13  
O3GEN_STATUS  
YES  
YES  
Switches ON/OFF power to the ozone generator.  
PREREACT_VALVE  
Switches the Pre-reactor valve. Use this bit to test the  
valve function.  
14  
15  
PREAMP_RANGE_HI  
ST_RCEL_PRESS  
YES  
YES  
Switches the preamp hardware range. Only the low  
range is active in the M200AU  
Status Bit - Reaction Cell Pressure alarm  
Logic High = pressure out of acceptable range  
Logic Low = pressure inside acceptable range  
Status Bit - Zero Calibration mode  
Logic High = M200AU in Zero cal mode  
Logic Low = Not in Zero cal mode  
Status Bit - Span Calibration mode  
Logic High = M200AU in Span cal mode  
Logic Low = Not in Span cal mode  
Status Bit - Flow alarm  
16  
17  
18  
19  
20  
21  
ST_ZERO_CAL  
YES  
YES  
YES  
YES  
YES  
YES  
ST_SPAN_CAL  
ST_FLOW_ALARM  
ST_TEMP_ALARM  
ST_DIAG_MODE  
ST_POWER_OK  
Logic High = Sample/Ozone flow out of spec  
Logic Low = Flows within spec  
Status Bit - Temperature alarm  
Logic High = Rxcell, Moly, Box temps out of spec  
Logic Low = Temps within spec  
Status Bit - In Diagnostic mode  
Logic High = M200AU in Diagnostic mode  
Logic Low = Not in Diag mode  
Status Bit - Power OK  
Logic High = Instrument power is on  
Logic Low = Instrument power is off  
(table continued)  
9-14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 9-4: Diagnostic Mode - Signal I/O (Continued)  
No.  
Signal  
Control  
Description  
22  
ST_SYSTEM_OK  
YES  
Status Bit - System OK  
Logic High = No instrument warnings present  
Logic Low = 1 or more alarms present  
Status Bit - Autorange High Range  
23  
ST_HIGH_RANGE  
YES  
Logic High = M200AU in high range of autorange  
mode  
Logic Low = M200AU in low range of autorange  
mode  
24  
PMT_SIGNAL  
NO  
Current PMT voltage. Same as PMT voltage in TEST  
menu. Bi-polar, typically in 0-5000 mV range. A  
constant value of 5000 mV indicates offscale.  
25  
26  
SAMPLE_PRES  
RCELL_TEMP  
NO  
NO  
Sample pressure in mV. Typical sea level value =  
4300 mV for 29.9" Hg-A.  
Reaction Cell temperature. Typically 3500 mV for  
50° C.  
27  
28  
29  
BOX_TEMP  
BLOCK_TEMP  
PMT_TEMP  
NO  
NO  
NO  
Box Temperature. Typically 1800 mV for 25° C  
Block temperature.  
PMT cold block temperature. Typically 3600 mV for  
10° C.  
30  
31  
DCPS_VOLTAGE  
RCELL_PRESS  
NO  
NO  
DC power supply composite voltage output.  
Typically 2500 mV.  
Reaction Cell Pressure in mV. Typically 1270 mV  
for 5" Hg-A at sea level. Is an absolute pressure so  
higher values means higher absolute pressures.  
32  
33  
34  
OZONE_FLOW  
CONV_TEMP  
NO  
NO  
NO  
Ozone flowmeter voltage. Typically 2000 mV at  
80 cc/min.  
Molybdenum Converter temp. Typically 3150 mV at  
315° C  
HVPS_VOLTAGE  
HVPS programming voltage. Output of HVPS is  
1000x value present. 700 mV = 700 VDC output.  
35  
36  
DAC_CHAN_0  
DAC_CHAN_1  
NO  
NO  
Output of NOx channel in mV.  
Output of NO channel in mV.  
(table continued)  
9-15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 9-4: Diagnostic Mode - Signal I/O (Continued)  
No.  
37  
Signal  
Control  
NO  
Description  
DAC_CHAN_2  
DAC_CHAN_3  
NOX_CONC  
Output of NO2 channel in mV.  
Test Channel output.  
38  
NO  
39  
YES  
NOx DAC programming voltage. The following 4  
signals can be set to output specific voltages to each  
DAC. Use in conjunction with ANALOG OUTPUT  
test to check each DAC output channel. The value  
keyed in should appear on the appropriate analog  
output channel. This value overrides data being  
written from the analyzer. Value reverts to instrument  
output when function is exited.  
40  
41  
42  
NO_CONC  
YES  
YES  
YES  
NO DAC programming voltage. See above for  
description.  
NO2_CONC  
TEST_OUTPUT  
NO2 DAC programming voltage. See above for  
description.  
TEST channel programming voltage. See above for  
description.  
9.1.3.2 Electric Test  
This function injects a constant voltage between the preamplifier and the buffer amplifier on the  
preamp board. Electric test checks part of the preamp, the V/F and computer for proper  
functioning. The result of electric test should be a smooth quiet signal as shown by constant  
values for the NO, NOx concentrations, the NO2 concentration should be near zero. Likewise the  
analog outputs should produce a smooth quiet trace on a strip chart.  
Procedure:  
1.  
2.  
Scroll the TEST function to PMT.  
Press SETUP-MORE-DIAG, then scroll to ELECT TEST by pressing the NEXT button.  
When ET appears, press ENTR to turn it on.  
3.  
The value in PMT should come up to 2000 mV ± 500 mV in less than 15 sec.  
If the HVPS or the span gain adjust on the preamp card has been changed without doing the  
FACTORY CALIBRATION procedure, the reading in step 3 may be different than 2000  
mV, since the overall calibration affects ELECTRIC TEST. See Section 9.1.6 for factory  
calibration procedure.  
9-16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
4.  
To turn off ET press EXIT  
If ET is a steady 2000 ± 500 mV, that means the Power Supply Module, Preamp buffer  
amplifier, V/F, CPU, and display are all working properly.  
9.1.3.3 Optic Test  
Optic test turns on a small LED inside the PMT housing which simulates the signal from the  
reaction cell. OT tests the entire signal detection subsystem. By observing the level, noise and  
drift of this test, correct operation of many sections of the analyzer can be verified.  
The implementation of OT involves several changes to the instrument operating conditions. The  
M200AU does the following when switching to optic test:  
1.  
2.  
Saves the current instrument setup as to autorange, indep range, current range and places  
the instrument into the 2000 ppb range.  
Turns off power to the ozone generator to assure there is no interfering light from the  
reaction cell.  
3.  
4.  
Disables the PRE-REACTOR circuit.  
Turns on the OT LED.  
Procedure:  
1.  
2.  
Scroll the TEST function to PMT.  
Press SETUP-MORE-DIAG, then scroll to OPTIC TEST by pressing the NEXT button.  
When OT appears, press ENTR to turn it on.  
3.  
4.  
The value in PMT should come up to 2000 mV ±1000 mV in less than 15 sec.  
To turn off OT press EXIT.  
If the PMT reading observed in step three comes up to the stated value, the instrument is capable  
of responding to the light generated by the chemiluminescent reaction. If the instrument passes  
OT, but there is absolutely no response to span gas, the most common cause is that the ozone  
generator is either broken or is turned off. A second common cause, is that what is thought to be  
span gas is actually zero gas.  
If the HVPS or the span gain adjust on the preamp card has been changed without doing a  
FACTORY CALIBRATE the reading in step 3 may be different than 2000 mV, since the overall  
calibration affects OPTIC TEST. See Section 9.1.6 for factory calibration procedure.  
9-17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
9.1.3.4 Ozone Gen Power  
This diagnostic manually turns the power off and on to the ozone generator. When the M200AU  
is powered up from a cold start the ozone generator is not immediately started. This is due to the  
fact that humid air may be present in the generator cartridge. Humid air can produce nitric acid  
aerosol which can permanently damage parts of the instrument down stream of the generator.  
Using this diagnostic, it is possible to turn on the generator before the warmup time has elapsed.  
If you turn the power on it will remain on after you exit the diagnostic.  
9.1.3.5 Analog Out Step Test  
The Step Test is used to test the functioning of the 4 DAC outputs on the V/F board. The test  
consists of stepping each analog output 0-20-40-60-80-100% of the output. If the analog outputs  
are set for 0-5V full scale the outputs would step 0-1-2-3-4-5 VDC. The stepping can be halted at  
any value by pressing the key under the percentage on the front panel. When the test is halted,  
square brackets are placed around the percentage value in the display. Pressing the key again  
resumes the test. This test is useful for testing the accuracy/linearity of the analog outputs.  
9.1.3.6 DAC Calibration  
The Digital to Analog Converters (DAC) are calibrated when the instrument is set up at the  
factory. Re-calibration is usually not necessary, but is provided here in case the V/F board needs  
to be replaced and re-calibrated. The procedure for doing the DAC Calibration routine is in the  
Troubleshooting Section 9.3.3.1.  
9.1.3.7 RS-232 Port Test  
This test is used to verify the operation of the RS-232 port. It outputs a 1 second burst of the  
ASCII letter 'w'. During the test it should be possible to detect the presence of the signal with a  
DVM on pin 2 or 3 (depending on the DTE/DCE switch setting) or by the flickering of the red  
test LED. A detailed procedure is given in the Troubleshooting Section 9.3.2.  
9-18  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
9.1.4 M200AU Internal Variables  
The M200AU software contains many adjustable parameters. Most of the parameters are set at  
time of manufacture and do not need to be adjusted for the lifetime of the instrument. Some of  
the variables are user adjustable, they are listed in Table 9-5.  
To access the VARS menu press SETUP-MORE-VARS-ENTR. Use the PREV-NEXT buttons  
to select the variable of interest, then press EDIT to examine/change the value, then press ENTR  
to save the new value and return to the next higher menu. If no change is required, press EXIT.  
TPC_ENABLE  
The M200AU has temperature and pressure compensation. T/P comp adjusts the output of the  
instrument for changes in sample temperature, reaction cell pressure, and atmospheric pressure.  
The temperature of the reaction cell controls the sample temperature. The setpoint is 40° C, and  
the value of the adjustment parameter is equal to 1.0000 when the reaction cell temperature is  
40° C. The temperature compensation increases sample concentration with increasing  
temperature to compensate for the drop in density of gas in the reaction cell.  
The reaction cell pressure compensation factor is equal to 1.0000 when the cell pressure is about  
3.5"-Hg-A. The compensation factor increases sample concentration with increasing cell  
pressure to compensate for increased quenching of the chemilumenscent reaction at higher  
pressures.  
The sample pressure compensation factor is equal to 1.0 at 29.92"-Hg-A. This factor increases  
sample concentration with decreasing sample pressure to compensate for a lower head pressure  
on the sample flow orifice.  
Taken together, the three factors change the output of the instrument very little. The sample  
temperature is essentially invariant, and the cell pressure and sample pressure factors tend to  
cancel each other. The resultant coefficient has no practical variation with pressure changes due  
to weather fronts. Changes in altitude of 1000 feet usually change the output of the instrument by  
about .5% if compensation is turned off, much less if it is operating.  
9-19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 9-5: Model 200AU Variables  
Default  
Value  
Value  
Range  
No.  
Name  
Units  
Description  
0
DAS_HOLD_OFF  
Min  
15  
0-60  
Time that data is not put into  
DAS after CAL or DIAG  
modes  
1
MEASURE_MODE  
Logic  
NONOX  
Selects the measurement  
mode of the M200AU. NO  
measures NO only. NOx  
measures NOx only, and  
NONOx measures NO, NOx  
and NO2, this is the default  
mode.  
2
3
4
TPC_ENABLE  
DYN_ZERO  
DYN_SPAN  
Logic  
Logic  
Logic  
ON  
ON-OFF  
ON-OFF  
ON-OFF  
Temp/Pres compensation  
enable  
OFF  
OFF  
Enable zero calibration when  
rear panel contacts are closed.  
Enable span calibration when  
rear panel contacts are closed.  
5
6
7
SFLOW_SET  
OFLOW_SET  
RS232_MODE  
cc/min. 1000  
cc/min. 80  
400-1200  
0-500  
Nominal sample flow rate  
Nominal ozone flow rate  
Bit  
0
0-99999  
Value is SUM of following  
decimal numbers:  
Field  
1=enable quiet mode  
2=enable computer mode  
4=enable security feature  
8=enable front panel RS-232  
menus  
16=enable alternate protocol  
32=enable multidrop protocol  
8
CLOCK_ADJ  
Sec.  
0
Real-time clock speed  
adjustment  
± 60  
9-20  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
9.1.5 Test Channel Analog Output  
Many of the TEST functions have an analog voltage associated with them. As a diagnostic aid it  
is possible to route the various test voltages out the 4th analog output port. Details on using the  
test channel analog output are in the Troubleshooting Section.  
Table 9-6: Test Channel Readings  
TEST Channel  
Minimum*  
Maximum*  
Description  
PMT  
DETECTOR  
0 mV  
5000 mV  
PMT detector output from the preamp. This  
signal has been amplified and filtered. Since the  
instrument is switched and uses AutoZero,  
normal values can vary from –100 to 5000 mV.  
Wide variations in this signal are normal.  
Values should be around 0 mV when sampling  
zero air.  
OZONE FLOW  
0 cc/min  
1000 cc/min  
This signal is the output from the ozone  
flowmeter. Values around 1150 mV indicate  
zero flow. Typical values for 80cc/min ozone  
flow are around 1800 mV. Voltage should be  
steady, indicating stable flow.  
SAMPLE  
FLOW  
0 cc/min  
1000 cc/min  
40 “-Hg-Abs  
The sample flow is calculated from the  
upstream pressure as measured by the  
SAMPLE PRESSURE transducer.  
SAMPLE  
PRESSURE  
0 “ Hg-Abs  
The sample pressure is measured by an absolute  
pressure meter. The absolute pressure at sea  
level is 29.92”-Hg. The exact reading will vary  
a few tenths due to passing weather fronts and  
daily temperature cycling. The reading will  
decrease about 1”-Hg with each 1000 ft gain in  
altitude. For example, the absolute pressure at  
10,000-ft (3000 m) is about 20”-Hg-A. A  
typical value near sea level would be about  
4200 mV.  
RCELL  
PRESSURE  
0 “ Hg-Abs  
40 “-Hg-Abs  
Like the SAMPLE PRESSURE the RCELL  
pressure is an absolute pressure measurement.  
With the sample pump off, it should read about  
atmospheric pressure. With the pump  
operating, a typical value is 1300 mV for about  
5”-Hg-A reaction cell pressure.  
(table continued)  
9-21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 9-6: Test Channel Readings (Continued)  
TEST Channel  
Minimum*  
Maximum*  
Description  
RCELL TEMP  
0o C  
70o C  
Reaction Cell temperature is set to 50o C. At the  
setpoint, a typical reading is 3600 mV.  
BLOCK TEMP  
IZS TEMP  
0o C  
0o C  
0o C  
0o C  
70o C  
The Block temperature is set to 50o C. At the  
setpoint, a typical reading is 3600 mV.  
The IZS temperature is set to 50o C. At the  
setpoint, a typical reading is 3600 mV.  
The Converter temperature is 315o C. At the  
setpoint, a typical voltage is 3150 mV.  
70o C  
CONV TEMP  
PMT TEMP  
1000o C  
70o C  
The PMT temperature is unique in that the  
voltage is inverse to the temperature. A typical  
reading for 8o C would be 4200 mV.  
CHASSIS  
TEMP  
0o C  
70o C  
The Chassis (Box) temperature is variable due to  
variable ambient air temperature. The Box temp  
generally runs about 5o C above the surrounding  
air temp. Thus in a 25o C room, the Box temp  
would be about 30o C and have a TEST channel  
voltage of about 2000 mV.  
DCPS  
VOLTAGE  
0 mV  
0 V  
5000 mV  
5000 V  
The DCPS is a composite of several DC power  
supply voltages in the instrument. It has been  
arbitrarily set at 2500 mV, which is typical.  
HVPS  
VOLTAGE  
The HVPS voltage is a scaled up reading of the  
programming voltage going to the HVPS. Zero  
to 1000 mV corresponds 0-1000 VDC for the  
HVPS, which is the maximum voltage possible.  
A typical reading would be 700 mV  
corresponding to 700 VDC for the HVPS.  
* Minimum and Maximum readings depend on the DAC 3 switch settings of the V/F board. For  
the standard ± 5 VDC range, minimum corresponds to 0 VDC and maximum corresponds to  
5 VDC.  
9-22  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
9.1.6 Factory Calibration Procedure  
This procedure is similar to that used at the factory when the instrument is first manufactured, it  
balances the PMT, preamp, and software gain factors so the instrument has optimum noise,  
linearity, and dynamic range. It should be used when you are unable to zero or span the  
instrument, when the slope and offset values are outside of the acceptable range, or when other  
more obvious reasons for problems have been eliminated.  
Factory Calibration Procedure:  
NOTE  
In this procedure a range of 500 ppb and a span gas concentration  
of 400 ppb is used as an example. Other values can be used.  
1.  
2.  
On the Preamp board, set S1 and S2 to 8. Turn R19 25 turns counter-clockwise, then 12  
turns clockwise.  
Set RANGE MODE to SNGL by SETUP-RNGE-MODE-SNGL to select single range  
operation.  
3.  
4.  
Set the RANGE to 500 ppb by SETUP-RNGE-SET and key in 500, then press ENTR.  
Input Zero gas into the sample port, and Scroll to the TEST function labeled PMT.  
Typical reading should be less than 50 mV. Readings above 150 mV indicate a pneumatic  
leak, light leak, contaminated reaction cell, bad zero gas, or wet air coming into the ozone  
generator. If readings are greater than ± 150 mV, the instrument will not zero or span  
properly, see Sections 9.2.8, 9.2.9.  
5.  
6.  
Allow the instrument to sample zero gas for at least 20 minutes to re-fill the internal data  
filters and Pre-reactor filter with zero readings. Then zero the instrument by CAL-ZERO-  
ENTR.  
Set the expected span concentration to 400ppb. Enter the expected NOx concentration of  
400 ppb by pressing CAL-CONC-NOX. Then press CAL-CONC-NO, to enter the expected  
NO concentration of 400 ppb. Then press EXIT to return to the CAL menu.  
7.  
8.  
Input 400 ppb of NO span gas in the sample inlet port.  
Scroll to the NORM_PMT - TEST function.  
9-23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
9.  
Calculate the expected NORM_PMT mV reading.  
Multiply the expected span value by 2 to get the mV reading. In this example the expected  
span gas concentration is 400 ppb and therefore the expected voltage is 800 mV. As an  
alternate method, the voltage can be determined from the graph in Figure 9-1. On the Y-axis  
find the calibration concentration in ppb, then determine the expected voltage from the X-  
axis.  
10.  
11.  
Adjust S2, the HVPS coarse adjustment, on the preamp board to the setting that produces  
a signal that is closest to 800 mV. Adjust S1, the HVPS fine adjustment, to the setting that  
produces a signal that is closest to 800 mV. Use R19 to trim the reading to 800 ± 50 mV. The  
readings will periodically go to zero as the Pre-reactor circuit operates, ignore the zero  
readings.  
Allow the instrument to sample span gas for 30 minutes. Then do a span calibration by  
CAL-SPAN-ENTR. After the span is completed, do the span quality check procedure in  
Table 7-18. This procedure is extremely important to assure that the instrument will operate  
with optimum noise, linearity, and dynamic range.  
12. Electric Test (ET) Procedure:  
A. To adjust ET press SETUP-MORE-DIAG, then scroll to ELEC TEST and press ENTR.  
B. Scroll the TEST functions until PMT is displayed.  
C. Adjust R27 until 2000 mV ± 200 is displayed.  
D. Press EXIT to return to SAMPLE mode.  
13. Optic Test (OT) Procedure:  
A. To adjust OT press SETUP-MORE-DIAG, then scroll to OPTIC TEST and press ENTR.  
B. Scroll the TEST functions until PMT is displayed.  
C. Adjust R25 until 2000 mV ± 250 is displayed.  
D. Press EXIT to return to SAMPLE mode.  
If this procedure does not produce an instrument that will properly span, please contact your  
local distributor or the Teledyne API factory. Teledyne API's phone number is on the front page  
of this manual.  
9-24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Figure 9-1: Span Calibration Voltage  
9-25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
9.2 Performance Problems  
When the response from a span check is outside the control limits, the cause for the drift should  
be determined, and corrective action should be taken. Some of the causes for drift are listed  
below:  
NOTE  
It has been our experience that about 50% of all analyzer performance  
problems are sooner or later traced to leaks in some part of the system.  
1.  
2.  
3.  
Fluctuations in flow. Such as leaks or plugged orifices.  
Lack of preventive maintenance.  
Change in zero air source.  
A. Air containing NO leaking into zero air line.  
B. Saturation of charcoal and/or Purafil scrubbers.  
Change in span gas concentration.  
4.  
A. Zero air or ambient air leaking into span gas line.  
B. Permeation tube or cal gas tank exhaustion.  
Leak in NO/NOx or Pre-reactor switching valves.  
Loose pneumatic fittings.  
5.  
6.  
9.2.1 AC Power Check  
1.  
Check that power is present at main line power input. Verify that correct voltage and  
frequency is present. If unit is set for 240 VAC and is plugged into 115 VAC it will appear as  
no power fault.  
2.  
3.  
Check that the unit is plugged into a good socket. Analyzer must have 3-wire safety  
power input.  
Check circuit breaker. Circuit breaker is part of the front panel power switch. It is set  
each time the instrument power is turned on. If there is an internal short causing a trip, the  
switch will automatically return to the OFF position when an attempt is made to turn it on.  
9-26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
9.2.2 Flow Check  
1.  
Check TEST function RCEL - this is the absolute pressure in the reaction cell. It should  
be  
1-4 in-Hg-A.  
2.  
3.  
4.  
Check that pump is running. Check RCEL - TEST function for proper pressure.  
Check that pump tubing is connected to rear of analyzer.  
Test that the pump is producing vacuum by removing fitting at rear of analyzer and  
checking for suction at fitting.  
5.  
Check for flow upstream of the sample flow orifice and ozone orifice on the valve  
assembly.  
A. Remove the 1/8" fitting that carries sample. Plug the fitting on the sample flow block  
with your finger and note the vacuum produced.  
B. Remove the ozone fitting also and compare relative flow rates. Sample should be much  
higher (1000 cc/min) than ozone (80 cc/min).  
6.  
7.  
Check for broken flow or pressure sensor.  
Leak check the analyzer. See Section 8.7 for leak check procedure.  
9.2.3 No Response to Sample Gas  
1.  
Confirm correct operation of analog output by performing Analog Output Step Test in  
Section 9.1.3.5.  
2.  
Confirm general operation of analyzer.  
A. Check for AC Power, Section 9.2.1.  
B. Do flow checks, Section 9.2.2.  
C. Confirm that sample gas contains NO or NO2.  
Check instrument electronics.  
3.  
A. Do ELEC TEST procedure in DIAGNOSTIC menu Section 9.1.3.2.  
B. Do OPTIC TEST procedure in the DIAGNOSTIC menu Section 9.1.3.3.  
C. If the M200AU passes ET and OT, that means the instrument is capable of detecting light  
and processing the signal to produce a reading. Therefore, the problem is in the  
pneumatics or ozone generator.  
4.  
5.  
Check ozone generator subsystem. Do the diagnostic test of the ozone generator  
subsystem in Section 9.3.6.  
Check for disconnected electrical cables to sensor module.  
9-27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
9.2.4 Negative Output  
1. Mis-calibration. The 'zero' gas that was used to zero the M200AU contained some NO gas -  
that is, it had more NO gas than that of the sample air. May also be caused by doing a zero  
calibration using ambient air. If NO/NOx OFFSET - TEST functions are greater than 150  
mV, reaction cell contamination is indicated. Verify quality of zero air used it should have  
less than 1 ppt NO and correspondingly low equivalent values of interferent gasses .  
2. Corruption of the Pre-reactor filter. If a significant signal was detected during the Pre-reactor  
cycle, that higher reading can enter the Pre-reactor filter. The value of the Pre-reactor filter is  
subtracted from the current reading, thus producing a negative reading. High Pre-reactor  
readings can be caused by:  
A. Leaking Pre-reactor valve.  
B. Electronic fault in the preamp causing it to have a voltage on the PMT output pin during  
the Pre-reactor cycle. Area around the high impedance section of the preamp contains  
conductive dirt.  
C. Reaction cell contamination causing high background ( >40 mV) light readings.  
D. Broken PMT temperature control circuit, or broken PMT cooler power supply - allowing  
high zero offset.  
After fixing the cause of the high Pre-reactor filter readings, the M200AU will take 15  
minutes for the filter to clear itself.  
3. Check for leaks.  
9.2.5 Excessive Noise  
Common reasons for excessive noise are:  
1.  
2.  
3.  
4.  
Leak in pneumatic system.  
Light leak - check the sensor module with strong light.  
HVPS noisy - see HVPS test procedure. See Section 9.3.8.5.  
Defective electronic components on preamp board. - use optic test and electric test to  
check electronics, optics and observe noise.  
5.  
Contamination of ozone generator and/or reaction cell - This can be wet air or impurities.  
This can be detected by high PMT readings with zero air as sample gas. Verify this condition  
by turning off the ozone generator using the DIAG mode command and observing a drop in  
PMT reading of more than 50 mV or 25 ppb. If the ozone generator or reaction cell is  
contaminated, disassemble and clean.  
9-28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
6.  
Broken PMT temperature control circuit or broken PMT cooler power supply. Check  
PMT TEMP - TEST function.  
7.  
8.  
Mis-calibration. Check NO/NOx SLOPES in TEST function.  
Reaction cell pressure too high – leak on vacuum side or bad sample pump. Check RCEL  
- TEST function.  
9.2.6 Unstable Span  
Common causes are:  
1.  
2.  
3.  
4.  
5.  
6.  
7.  
8.  
Leak in pneumatic system.  
Light leak - check the sensor module with strong light.  
Instrument not fully warmed up.  
Sample lines or sample filter dirty - clean or replace.  
Plugged sample inlet orifice - clean with methanol and sonic cleaner.  
Defective HVPS - see HVPS test procedure.  
Bad or defective PMT detector - replace.  
Reaction cell temp not stable - observe warning messages, or RCELL TEMP in TEST  
functions. Check diagnostic LED in Power Supply Module for normal cycling.  
9.  
Large variations in ambient temperature - observe warning messages, or BOX TEMP in  
TEST functions.  
10.  
11.  
Large variations in line voltage. - Line voltage should remain within ±10% of nominal.  
Pump not maintaining steady vacuum - observe warning messages, or RCEL in TEST  
functions.  
12.  
13.  
Sample vent line too short, allowing room air to mix with span gas. Line should be a  
minimum of 15" long.  
Calibration gas source unstable.  
9-29  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
9.2.7 Unstable Zero  
Common causes are:  
1.  
2.  
3.  
4.  
5.  
Leak in pneumatic system. Perform leak check, see Section 8.7.  
Miscalibration. See Table 7-1.  
Light leak - check the sensor module with strong light.  
Sample lines or sample filter dirty - clean or replace.  
PMT temp control unstable  
9.2.8 Inability to Span  
If the SPAN button is not illuminated when attempting to span, that means the reading is outside  
of the software gain ranges allowed. In an analog instrument it would be the equivalent to the  
span pot hitting the rotation stop.  
Here are some things to check:  
1.  
Check the expected span concentration values in CAL-CONC-NOX and CAL-CONC-  
NO, and compare them to the values of the calibration span gas being input. They should be  
nearly equal.  
2.  
3.  
Check the NORM_PMT - TEST function. With NO span gas in the instrument, the value  
should be 2x the expected span concentration.  
Check ET and OT to verify instrument detector and signal processing is working  
correctly.  
4.  
5.  
If the above do not check out, perform the Factory Calibration Procedure Section 9.1.6.  
If the PMT voltage is near zero with span gas, check fuse and power to ozone generator.  
9-30  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
9.2.9 Inability to Zero  
If the ZERO button is not illuminated when attempting to zero, that means the reading is outside  
of the software gain ranges allowed. In an analog instrument it would be the equivalent to the  
zero pot hitting the rotation stop. Check the following:  
Select the PMT - TEST function. With zero gas going into the instrument, the value should be  
less than 150 mV, typically less than 50 mV. If you are getting a high reading, the probable  
reasons are:  
1.  
The reading may be temporarily high if the PMT has been recently exposed to room  
light. If so, let the instrument run for several hours with zero gas to get the PMT accustomed  
to low light levels.  
2.  
3.  
4.  
Leak that admits gas containing NO into the reaction cell.  
Contaminated reaction cell. Remove and clean cell.  
Wet (i.e., undried ambient air) air in the ozone generator. Check the PermaPure drier and  
associated plumbing for leaks and correct operation.  
5.  
6.  
Zero gas that isn't really zero. Make sure you're not trying to zero the machine with  
ambient air or span gas.  
Pre-reactor filter is corrupted with high readings. To clear the Pre-reactor filter, input  
zero gas and wait for 15 minutes for the filter to clear.  
9.2.10 Non-Linear Response  
Common causes are:  
1.  
2.  
A leak in the pneumatic system, see Section 8.7.  
High zero background - the PMT TEST function should be near 50 mV with zero gas.  
Readings above 100 mV indicate a light leak, contaminated reaction cell, bad zero gas, or  
wet air coming into the ozone generator. If the reading is not less than 150 mV, the  
instrument will not zero or span properly. Check Section 9.2.9.  
3.  
Calibration device in error - re-check flowrates and concentrations, especially at low  
concentrations. If you are using a Mass Flow calibrator and the flow is < 10% of the full  
scale flow on either flowmeter you may need to purchase lower concentration standards.  
4.  
5.  
The standard gasses may be mis-labeled as to type or concentration. Labeled  
concentrations may be outside the certified tolerance.  
Contamination in sample delivery system:  
A. Dirt in sample lines or reaction cell  
B. Contaminated cal gas source (NO2 in NO cal gas is common)  
C. Dilution air contains sample or span gas  
9-31  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
6.  
Ozone concentration too low  
A. Wet air in ozone generator - need to flush out clean, dry  
B. Ozone generator driver board failure.  
C. Damaged transformer/ozone generator cartridge. The ozone generator cartridge is  
partially made of glass. Extreme physical shock, such as dropping the generator on the  
floor may cause the glass cartridge to crack and eventually fail.Sample inlet vent line too  
short - should be at least 15".  
7.  
Sample exhaust not properly vented, creating a backpressure at the sample inlet port of  
the instrument. See Figure 2-3 for venting recommendations.  
9.2.11 Slow Response  
1.  
Contaminated or dirty sample delivery pneumatics  
A. Dirty/plugged sample filter or sample lines  
B. Dirty reaction cell  
2.  
3.  
4.  
Sample inlet line too long.  
Wrong materials in contact with sample - use glass, stainless steel or Teflon  
Sample vent line located too far from instrument sample inlet. Locate sample inlet vent  
near analyzer.  
5.  
6.  
7.  
8.  
9.  
Insufficient time allowed for purging of lines upstream of analyzer.  
Leaking NO/NOx valve.  
Insufficient time allowed for NO or NO2 cal gas source to become stable.  
Moly converter temperature too low.  
Miscalibration, see Table 7-1.  
9.2.12 Analog Output Doesn't Agree with Display Concentration  
The analog output is proportional to the range. Zero volts output corresponds to zero ppb, and 5  
volts corresponds to the maximum range setting in ppb. If this is not observed do the following:  
1.  
2.  
V/F board DAC's out of calibration. Do DAC calibration and Factory Calibration.  
Analog outputs electrically loaded down causing voltage to sag. Could be due to input  
impedance to chart recorder or data logger being too low or improper grounding. The  
Recorder and DAS outputs do not have separate output drivers, the problem could be the  
combined load of both could be too high.  
9.3 Subsystem Troubleshooting and Adjustments  
9.3.1 Computer, Display, Keyboard  
9-32  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
The purpose of this section is to determine if the computer subsystem electronics hardware are  
working properly. Assessment will be made at the board level.  
9.3.1.1 Front Panel Display  
The front panel display is a 2 line by 40 character display. It has its own microprocessor to  
decode commands and display characters. It contains a self test feature. To test the display:  
1.  
2.  
3.  
Turn off the power to the instrument.  
Fold down the M200AU front panel  
Disconnect the 26 line flat ribbon cable (J2) that connects the computer parallel port to  
the keyboard.  
4.  
5.  
Turn on the M200AU power switch.  
Observe the front panel display. If the display successfully completes its power on self  
test, it will display a single underline character "_" in the left most character of the top line of  
the display. If this character is present, the display is working properly.  
6.  
Power down the analyzer, and re-attach the cable to J2, and proceed to the next test.  
9.3.1.2 Single Board Computer  
The SBC40 is a full function computer designed for instrument control applications. It consists  
of a 16 bit V40 microprocessor, 2 serial and one parallel ports, Standard Bus interface, and 4  
sockets for memory. The memory sockets consist of: 256k ROM containing the multitasking  
operating system and application code, 8k EE prom containing the setup variables, a 128k RAM  
and time-of-day clock to provide event timing services, and a second 128k RAM. The overall  
function of this board is quite complex, therefore, complete testing is not possible in the field.  
9-33  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Figure 9-2: CPU Board Jumper Settings  
9-34  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Like the display, the overall functioning of the CPU can be confirmed by a simple test.  
1.  
2.  
3.  
4.  
5.  
Locate the CPU board on the mother board by referring to Figure 2-5.  
Power the instrument on.  
Locate the red LED at the top left edge of the board.  
It should be flashing at a frequency of about once per second.  
This flashing indicates the board is powered up and is executing instructions.  
Testing and operation of the CPU RS-232 port is described in Section 9.3.2. It is possible for the  
UART driver chip to malfunction in either the input or output port or both.  
9.3.1.3 Front Panel Keyboard  
The keyboard consists of 8 keys and 3 LED's. Key strokes are sent to the SBC40 computer's  
parallel port. The computer software detects the key strokes via interrupts. The bottom line of the  
display consists of 40 characters which is divided into 8 - 5 character fields. Each field defines  
the function of the key immediately below it. The definition of the keys is variable and depends  
on the menu level of the software.  
To check the operation of the keyboard, each key should perform an operation indicated by its  
current definition shown on the second line of the display.  
Example #1 - testing key#1 (left most key).  
At the top level menu key #1 is defined as the TEST function. Pressing this key should cause the  
middle field of the top line of the display to show the various test functions.  
Example #2 - testing key #8 (right most key). At the top level menu key #8 is defined as the  
SETUP key. pressing key #8 should cause the SETUP menu to be displayed.  
Example #3 - If the 5 character field above any key is blank, the key is not defined, pressing the  
key has no effect.  
The 3 status LED's indicate several functional states of the instrument such as calibration, fault,  
and sample modes. The state of the LED's is controlled by 3 lines on the parallel port of the  
SBC40. Functioning of the LED's can be checked by:  
1.  
2.  
3.  
Turn off the M200AU power.  
While watching the LED's, turn on the instrument power.  
When the power comes up, the computer momentarily applies power to all 3 LED's. If all  
the LED's are observed to light, they are working properly.  
9-35  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
9.3.2 RS-232 Communications  
The M200AU uses the RS-232 communications protocol to allow the instrument to be connected  
to a variety of computer based equipment. RS-232 has been used for many years and is well  
documented. Generally, every manufacturer observes the signal and timing requirements of the  
protocol very carefully. Problems arise when trying to specify connectors, and wiring diagrams  
that attach the analyzer to various devices.  
9.3.2.1 RS-232 Connection  
If the RS-232 port is not working, check the following:  
Physical Wiring  
First is to get the physical wiring hooked up correctly. Refer to Figure 9-3 for the wiring diagram  
of the DB-9 plug on the M200AU rear panel. There are 2 features that make connecting the  
wiring easier. First is the red/green LED’s on the rear panel. The M200AU provides the power to  
run the red LED, the external equipment will provide the power for the green LED. If the wiring  
is hooked up correctly both LED’s will be illuminated. Secondly, there is a DTE-DCE switch,  
this switch interchanges pin 2 & 3 on the DB-9 connector. Set the DTE-DCE switch so that both  
LED’s are illuminated.  
RS-232 Protocol (BAUD rate, Data bits, Parity)  
Second is to get the communication protocol for each instrument to match. In Figure 9-3 the  
default RS-232 parameters are listed. The BAUD rate can be changed in the software menus  
under SETUP-MORE-COMM-BAUD.  
Data Communications Software for a PC.  
You will need to have a software package to enable the computer to transmit and receive on its  
serial port. There are many such programs, we use PROCOMM at Teledyne API. Once you set  
up the variables in PROCOMM and your wiring connections are correct, you will be able to  
communicate with the analyzer.  
If connecting to a modem, check the following:  
Modems are especially difficult because they may have pins that need to be at certain EIA RS-  
232 levels before the modem will transmit data. The most common requirement is the Ready to  
Send (RTS) signal must be at logic high (+5V to +15V) before the modem will transmit. The  
Teledyne API analyzer sets pin 8 (RTS) to 10 volts to enable modem transmission.  
9-36  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
To troubleshoot a modem connection first disconnect the RS-232 cable from the Analyzer and  
verify (use a DVM) that you are getting a signal on Pin 2 of the RS232 port on the Analyzer. The  
signal will be between -5V and -15V with respect to signal ground (pin 5). If not, there is a  
problem with the CPU board or the cable. This is the transmit (TD) signal out of the Analyzer.  
This should then be connected to TD input on the modem, normally Pin 2. If the Analyzer is  
equipped with a DTE/DCE switch, you may need to change its setting so the signal is on Pin 2.  
Second: Go to the cable connected to the modem/terminal and verify (use a DVM) that you are  
getting a -5V to -15V signal on Pin 3 of the cable. This pin should be connected to Pin 3 of the  
Teledyne API Analyzer.  
Third: (for modems) Check that the voltage level on Pin 8 of the Analyzer is between +5V and  
+15V. This pin should be connected (through the cable) to Pin 4 of the modem.  
Now set the baud rate of the Analyzer to the speed required by the modem and it should work. If  
you are still experiencing problems, a cable adapter may be needed. Please contact the factory  
for assistance.  
9.3.2.2 RS-232 Diagnostic Procedures  
There are several features of the M200AU to make connecting to RS-232 and diagnosing RS-  
232 faults easier.  
There are two LED's on the rear panel Connector Board which are connected to pin 2 and 3 of  
the DB-9 connector on the board. If the switch is in the DCE position (default) the red LED is  
connected to pin 3 of the DB-9 connector. When data is transmitted by the M200AU the red  
LED will flicker, indicating data present on this line. When the M200AU is running, the LED  
will normally be ON, indicating logic low. A one second burst of data can be transmitted over  
the port by a command in the DIAGNOSTIC menu. Press SETUP-MORE-DIAG, then scroll to  
RS232 OUTPUT. Each time you press ENTR the instrument transmits a 1 sec burst of lower  
case "w"'s.  
The green LED is connected to pin 2, if the switch is in the default DCE position. This is the pin  
on which the M200AU receives data. The LED is ON if an outside device is connected. This  
LED gets its power from the outside device. When data is being transmitted by the outside  
device to the M200AU this LED will flicker.  
When you are attempting to configure the RS-232 port, if either of the LED's go out when the  
cable is connected, that generally means that there is a grounding problem. Switching the DCE-  
DTE switch should fix the problem. See the schematic and assembly drawings in the Appendix.  
9-37  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Figure 9-3: RS-232 PIN Assignments  
9-38  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
9.3.3 Voltage/Frequency (V/F) Board  
The V/F Board consists of 16 analog input channels, each software addressable, 8 digital inputs,  
and 24 digital outputs, each line independently addressable, and 4 independent analog output  
channels. The analog input channels are connected to V/F converter capable of 80,000 counts,  
which is approximately 16 bit resolution. The integration period is software selectable from  
40msec to 2.4 sec. Commands from the SBC40 computer and digitized values from the V/F  
section of the board are sent via the STD bus interface. The schematic for the board is in the  
Appendix.  
The overall operation of this board is quite complex. To fully check it out in all of its operational  
modes is not possible in the field. Therefore, a few simple tests are described here that test one  
analog input channel, the 4 analog output channels, one digital input, and one digital output.  
1.  
V/F board analog input test.  
Each analog channel is routed through a programmable 16 channel multiplexer. Chances are  
that if one channel works, they all work.  
A. Turn on instrument.  
B. Press TEST key on front panel keyboard until DCPS test is displayed.  
C. The value displayed should read 2500 ± 200 mV.  
If the M200AU passes this test, it has successfully digitized a 2500 mV composite voltage  
output from the Power Supply Module. The signal should also be quiet ±25 mV.  
2. Analog output channel test.  
In the DIAGNOSTIC menu on the front panel, there is a test that outputs a step voltage to the  
4 analog outputs. This test is useful for calibrating chart recorders and dataloggers attached  
to the M200AU. The test can also be useful in diagnosing faults in the V/F board.  
A. Turn on the instrument.  
B. Enter the SETUP-MORE-DIAG-ENTR menu.  
3.  
Select the ANALOG OUTPUT test. This causes the M200AU to output a 5 step voltage  
pattern to the 4 analog outputs on the rear panel. The status of the test is shown on the front  
panel display. The scrolling can be stopped at any voltage by pressing the key below the  
changing percentage display. The values are 0-20-40-60-80-100% of whatever voltage range  
has been selected. For example the voltages would be 0, 1, 2, 3, 4, 5V if the 5V range had  
been selected.  
9-39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
4.  
Use a DVM on each of the analog output channels to confirm the correct voltages.  
If the voltages step, but are the wrong values, the V/F board may be set up wrong or out of  
calibration. See Section 9.3.3.1 for information on how to calibrate the V/F board.  
5.  
Digital input channel test.  
The digital I/O section of the V/F board has 8 input bits and 24 output bits. Two of the 8  
input bits are assigned as calibration controls. See Section 7.5 for information on calibration  
using external contact closures.  
To test the digital inputs:  
A. Turn on the M200AU.  
B. Connect a jumper wire across pins 1 and 2 of the rear panel connector as shown in  
Figure 2-2.  
C. Shortly after closure is made the instrument should switch into zero mode as indicated on  
the front panel display.  
D. Remove the jumper.  
Digital output channel test.  
6.  
There are 24 output bits on the V/F board. The 24 bits are made up of 3 - 8 bit ports. It is  
possible for a single 8 bit port or even a single bit within a port to fail.  
A quick observational test of the digital outputs is to observe the LED's in the Power Supply  
Module(Refer to Figure 9-5 for the location of the LED's in the PSM) The state of the LED's  
can be checked from Table 4-1. The comments section assumes the M200AU has been  
running for at least 45 minutes.  
A more detailed test is in the DIAGNOSTIC menu. See Diagnostic tests in Section 9.1.3.  
9-40  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
9.3.3.1 ADC/DAC Calibration Procedure  
Due to the stability of modern electronics, this procedure should only need to be performed if a  
major sub-assembly is exchanged or when the display voltage does not match the input voltage  
or current to the V/F card. After completion, a Factory Calibration Procedure should be  
performed, see Section 9.1.6.  
Before the actual calibration is performed, switches on the V/F card must be correctly set and  
jumpers set on the motherboard. Jumper and switch setting changes must be performed with  
the instrument power OFF.  
Motherboard Jumpers  
The motherboard contains 4 pairs of jumpers JP1 - JP8, one pair for each analog output channel.  
Each channel can be configured for either voltage or current output. Use Table 9-7 to configure  
the jumpers.  
Table 9-7: Motherboard Jumper Settings  
Terminal Pair  
Rear panel  
Jumper  
Pair  
Jumper Setting for  
Voltage Mode  
Jumper Setting  
for Current Mode  
Analog output  
DAC 0 - NOx  
DAC 1 - NO  
DAC 2 - NO2  
DAC 3 - TEST  
3-4  
5-6  
1-2  
7-8  
JP3 - JP4  
JP1 - JP2  
JP5 - JP6  
JP7 - JP8  
B-C  
B-C  
B-C  
B-C  
A-B  
A-B  
A-B  
A-B  
V/F Board Switch Settings  
There are 2 different types of current outputs, Non-Isolated and Isolated. Each requires a different switch  
setting shown below. If you are operating the instrument in voltage output mode, the switches should be  
set to the desired voltage range.  
9-41  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 9-8: V/F Board Switch Settings  
DAC #  
Sw 1  
ON*  
ON*  
ON*  
ON*  
Sw 2  
Sw 3  
Sw 4  
Sw 5  
Sw 6  
Sw 7  
0
1
2
3
OFF*  
OFF*  
OFF*  
OFF*  
OFF*  
OFF*  
OFF*  
OFF*  
*Required settings  
10 V output or non-isolated current loop  
Switch  
State  
ON  
Comment  
3
4
5
6
Use for non-isolated current loop or 10 V output  
Use for non-isolated current loop or 10 V output  
Use for non-isolated current loop or 10 V output  
Use for non-isolated current loop or 10 V output  
OFF  
OFF  
OFF  
5 V output or isolated current loop  
Switch  
State  
OFF  
ON  
Comment  
3
4
5
6
Use for isolated current loop or 5 V output  
Use for isolated current loop or 5 V output  
Use for isolated current loop or 5 V output  
Use for isolated current loop or 5 V output  
OFF  
OFF  
1 V output  
Switch  
State  
OFF  
OFF  
ON  
Comment  
3
4
5
6
Use for 1 V output  
Use for 1 V output  
Use for 1 V output  
Use for 1 V output  
OFF  
100 mV output  
Switch  
State  
OFF  
OFF  
OFF  
ON  
Comment  
3
4
5
6
Use for 100 mV output  
Use for 100 mV output  
Use for 100 mV output  
Use for 100 mV output  
9-42  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
1.  
After the switches and jumpers are set, turn on instrument power and complete the  
following:  
A. Press SETUP-MORE-DIAG, then press ENTR. Scroll to D/A CALIBRATION, press  
ENTR.  
Press ADC to select the first task, which is to calibrate the A/D converter.  
B. Connect a DVM ground lead to TP3-AGND on the V/F board. Connect the positive lead  
to TP9-DAC0.  
C. The M200AU will display a voltage near 1% of the voltage range set in the above  
procedure. See Figure 9-4 for a table of approximate expected voltages. Adjust R27 until  
the displayed voltage matches the DVM voltage, then press ENTR.  
D. The M200AU will display a voltage near 90% of the voltage range set in the above  
procedure. Adjust R31 until the displayed voltage matches the DVM voltage, then press  
ENTR. This step calibrates the instrument A/D converter to the external DVM.  
E. The M200AU will automatically scroll through the DAC output channels, calibrating  
each one. The display will update the progress of the calibration. If operating all analog  
outputs in voltage mode this completes the calibration procedure.  
The calibration of each channel can be checked or recalibrated independently by scrolling the  
PREV-NEXT buttons and pressing CAL for the desired channel. Also, a DC offset bias can be  
entered by pressing the OFFSET button for the appropriate channel.  
Calibrating a channel for current loop operation:  
1.  
BEFORE STARTING, make sure that hardware settings are correct as shown in Table 9-  
6 and Table 9-7.  
2.  
3.  
Calibrate the A/D converter as described in item 2 above, if necessary.  
Connect a 250 ohm resistor in series with a current meter to the correct pair of terminals  
on the rear panel, see Figure 2-2 for terminal assignments.  
4.  
In the D/A calibration Menu press CFG, then use PREV-NEXT to select the channel to  
be calibrated.  
5.  
6.  
7.  
8.  
Press the SET button to select CURR for current loop operation, then press ENTR.  
To enter the calibration routine press CAL.  
To calibrate the 4 mA zero point, press the UP-UP10 - DN-DN10, then press ENTR.  
The M200AU will then output 20 mA. Press the UP-UP10 - DN-DN10 buttons to  
calibrate the span point, then press ENTR.  
9.  
Repeat steps 2-6 for each channel that needs to be calibrated.  
9-43  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Figure 9-4: V/F Board Settings  
9-44  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
9.3.3.2 Changing Output Voltage Ranges  
Several different output voltage ranges can be selected by switches on the V/F board. See Figure  
9-4 for the switch settings.  
9.3.4 Status/Temp Board  
The Status/Temp Board is a multi-function board that:  
1.  
2.  
3.  
4.  
5.  
6.  
Converts the resistance readings of the thermistors to voltages  
Conditions the thermocouple voltage for the V/F card  
Provides status output circuitry  
Provides circuitry for contact closure inputs  
Provides circuitry for display brown-out/reset at low line voltage  
Provides chip carriers for voltage-to-current modules  
9.3.4.1 Temperature Amplifier Section  
The Status/Temp board is a multi-function board consisting of 4 thermistor amplifiers that  
monitor:  
1.  
2.  
3.  
4.  
IZS temperature (not available on the M200AU)  
Reaction Cell temperature  
Box temperature  
Spare  
All 4 amplifiers have a single gain control - R34. If necessary, you can adjust the temperature  
values by selecting the BOX TEMP - TEST function and adjusting R34 until the correct box  
temp is shown. Remember that the temperature inside the case runs several degrees higher than  
room temperature, which should be taken into account when setting BOX TEMP.  
The voltages of the thermistor and thermocouple amplifier outputs are brought out to test points  
on the edge of the board. Refer to the Status/Temp board schematic for details. The voltages can  
also be read using the DIAGNOSTIC - SIGNAL I/O feature. See Table 9-4 for details.  
9-45  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
In addition there is a thermocouple amplifier for the Moly converter.  
Thermistor Temperature Amplifier Adjustments  
If the temperature readouts are in error:  
1.  
Locate the Box temp thermistor on the board and place a calibrated thermometer near the  
thermistor.  
2.  
3.  
Select the BOX TEMP - TEST function on the front panel.  
Adjust R34 until the front panel readout matches the thermometer readout. This will  
cause all of the readouts to accurately measure their respective temperatures.  
Molybdenum Converter Thermocouple Amplifier Adjustments  
The molybdenum converter temperature is sensed by a thermocouple. The cold junction  
compensation and signal conditioning is done on the Temp/Status board.  
The buffer amplifier from the thermocouple amplifier to the CPU for the Moly temperature has a  
gain adjustment. The voltage times 100 at pins 8-9 of U1 is the Moly temperature in degrees C.  
For example 3.15 VDC at pin 9 of U1 is 315o C. The CPU is programmed to always drive the  
temp to 315° C, so the voltage at U1 must be used as the absolute reference as to the correct  
temperature. The temperature will vary ± 3 degrees due to the operation of the temperature  
control loop in the CPU.  
The R6 pot on the Status/Temp board can adjust the temperature. To adjust the molybdenum  
converter temperature:  
1.  
2.  
3.  
Select the CONV TEMP - TEST function on the front panel.  
Wait until the converter is up to temperature, usually 30-45 min after a cold start.  
Adjust R6 until the voltage at pin 8-9 of U1 is 3.15 VDC. The CONV TEMP - TEST  
function will then read 315° C.  
4.  
Recheck the temperature 15 min later and re-adjust if necessary.  
9.3.4.2 Display Brownout  
During low AC line conditions the display can lock up due to insufficient voltage. When low  
line conditions are approaching, this circuit senses the condition by monitoring the un-regulated  
+5 VDC in the Power Supply Module. If brownout conditions are met, the DISP_BROWNOUT  
line is asserted and the CPU sends a hardware RESET command to the display and sends a  
BRNOUT RESET pulse back to U4. Brownout conditions will be noticed by the display flashing  
every 8 seconds.  
9-46  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
9.3.4.3 Status Output Lines, External Contact Closures  
The Status lines consist of 2 active input lines, and 12 active output lines. Additional circuits are  
present on the board but currently unused. Individual lines are set or cleared under CPU control  
depending on the assigned alarm condition. The CPU also monitors the 2 input lines for remote  
calibration commands. The status inputs and outputs are terminated at the rear panel, see Rear  
Panel Connector Board schematic diagram in the Appendix.  
The output lines are opto-coupled NPN transistors which can pass 50 ma max of direct current  
with a voltage of 30 VDC max.  
The input lines are optically coupled with inputs pulled up to +5 VDC. External contacts can be  
contact closures or open collector transistor contacts. DO NOT apply any voltage, since +5 VDC  
is supplied internally.  
Individual status lines can be set or cleared using the DIAGNOSTIC mode SIGNAL I/O. This  
can be useful for simulating fault conditions in the analyzer to see if external circuitry is working  
correctly. See Table 5-9 for pin assignments.  
9.3.4.4 4-20 mA Current Output  
4-20 mA current loop option replaces the voltage output of the instrument with an isolated 4-20  
mA current output. The current outputs come out on the same terminals that were used for  
voltage outputs, see Figure 2-2. It is programmable for 4-20mA or 0-20mA and has a 1500 V  
common mode voltage isolation and 240 V RMS normal mode voltage protection.  
Vloop = 28V max which is sufficient to drive up to a 1000 ohm load.  
9.3.5 Power Supply Module  
The Power Supply Module consists of several subassemblies described in Table 9-9.  
9-47  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 9-9: Power Supply Module Subassemblies  
Module  
Description  
Linear Power Supply  
Board  
The linear power supply board takes multiple voltage inputs from the  
power transformer and produces +5, +15, -15, +12 VDC outputs. The  
outputs are routed to two external connectors, P2 and P3. See  
Figure 9-5. The +5 is used for operating the CPU. The ± 15 is used in  
several locations for running op-amps and IC's. The +12 is used for  
operating fans and valves.  
Switching Power Supply  
Switch Board  
The switching power supply supplies +15 VDC at 4 A to the PMT  
cooler control on the Sensor Module. The output is made available  
through J10 on the Switch Board. There is a load resistor on the  
Switch Board to keep the output stable when little current is required  
from the supply.  
The Switch Board has many different functions. It takes logic signals  
from the V/F board and uses them to switch 4-115 VAC and 4-12VDC  
loads. The board also contains the instrument central grounding tie  
point. It routes unswitched AC and DC power as needed. Connector J2  
programs the power transformers to take 115, 220, and 240 VAC  
inputs  
Power Transformers  
There are potentially 2 input power transformers in the instrument. The  
multi-tap transformer T1 is in every M200AU and supplies input  
power for the Linear Power Supply board described above. A second  
transformer T2 is added if 220 or 240 VAC input is required. Input  
power selection is done via a programming connector P2 which  
provides the proper connections for either foreign or domestic power.  
Circuit Breaker/Power  
Switch  
The front panel contains a combination circuit breaker - input power  
switch. It is connected to the PSM through J6 on the Switch Board. If  
an overload is detected the switch goes to the OFF position. Switching  
the power back on resets the breaker also.  
9-48  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Figure 9-5: Power Supply Module Layout  
9-49  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Figure 9-6: Electrical Block Diagram  
9-50  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
PSM Diagnostic Procedures  
The Linear Power Supply board can be tested by checking the DCPS - TEST function on the  
front panel. It should read 2500 mV ± 200 mV. If the value is outside this range, individual  
output voltages can be tested on connector P3, see Schematic in the Appendix for pinouts.  
The Switching Power Supply output can be tested by observing the temperature of the PMT cold  
block using the PMT TEMP - TEST function. The temperature should be constant –5 ± 2°C. The  
output voltage can be observed on J10 of the Switch Board. It should be 15 VDC ± 0.5.  
The Switch Board can be tested by observing the diagnostic LEDS along the top edge of the  
board. The following Table 9-10 describes the typical operation of each LED.  
Table 9-10: Power Supply Module LED Operation  
No.  
Function  
Description  
1.  
NO/NOx Valve  
Should switch about every 5 sec.  
On(click sound) = NOx mode  
Off(thud sound) = NO mode  
2.  
Zero/Span Valve  
Should switch ON when CALZ or CALS button is  
pressed.  
3.  
4.  
Sample/Cal Valve  
Pre-reactor Valve  
Should switch ON when CALS button is pressed.  
ON when M200AU in Pre-reactor mode. Happens once  
every 6 NO/NOx cycles or about once per minute.  
5.  
Ozone Generator Power  
Ozone generator power is controlled by:  
1. Molybdenum converter temp  
2. Time since last power-up  
The ozone generator will be on if:  
1. Moly temp is < 100 C and its been more than 30 min  
since power up.  
2. Moly temp between 100-250 C and its been more than  
10 min since power up.  
3. Moly temp > 250 C  
6.  
7.  
8.  
Sample flow control block  
heater  
Should cycle ON-OFF every 20 sec to 2 min. On  
continuously until up to temp.  
Converter Heater  
Should cycle ON-OFF every 20 sec to 2 min. On  
continuously until up to temp.  
Reaction Cell Heater  
Should cycle ON-OFF every 20 sec to 2 min. On  
continuously until up to temp.  
9-51  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
9.3.6 Ozone Generator  
CAUTION - DANGER  
Lethal voltages present inside Ozone Generator Turn  
off instrument before servicing. Unplug Generator  
before dis-assembling.  
The ozone generator subsystem consists of a permeation drier, flowmeter, power supply -  
generator module, and power switch. The location of the components is illustrated in Figure 9-7.  
Ozone is generated by drying ambient air, passing the air between two electrodes that have a  
large oscillating electric field generated by a high voltage transformer.  
Common faults in the ozone generator are:  
1.  
A leak or some other failure in the drier will let ambient air into the generator. There is  
enough water vapor in room air to cause the generator to make nitric acid aerosol. It is very  
corrosive and causes the generator cartridge to short out due to salt build-up. This reduces  
the ozone concentration generated which can cause the analyzer to be non-linear due to  
insufficient ozone concentration.  
2.  
3.  
Contaminated ozone generator cartridge. Salts and oxgenated hydrocarbons can be  
removed by rinsing with 10% nitric acid at 50° C, followed by de-ionized water rinse, then  
dry air purge.  
Shorts, open circuits, arcing. Because of the very high voltages, the glass ozone generator  
cartridge may crack and cause an internal arc or short circuit. Measure the voltage between  
TP-4 and TP-5 on the Ozone Generator Driver Board. The voltage should be 0.045 ± .005  
VDC, indicating that the drive current is about .45 A. If the value is either zero or  
significantly greater than 0.45, the ozone generator is faulty.  
9-52  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 9-11: Ozone Generator Control Conditions  
Ozone  
Generator  
Condition  
Description  
1
2
3
OFF  
ON  
Manual override off (SETUP-MORE-DIAG-OZONE GEN).  
Manual override on.  
Ozone flow below low warning limit for 5 minutes (OFLOW_SET  
OFF  
setup variable, typically <40 cc/min).  
4
5
Instrument powered on for more than 30 minutes.  
ON  
ON  
Instrument powered off for less than 1 hour and ozone generator was  
on when instrument was powered off.  
6
Ozone flow above low warning limit for 0.5 minutes and condition 4  
ON  
or 5 is true.  
NOTE  
The ozone generator is independently controlled in the SIGNAL I/O, OPTIC  
TEST, and ELECTRICAL TEST diagnostics. After exiting these diagnostics,  
the ozone generator is restored to the state specified by the above conditions.  
9.3.6.1 PermaPure Drier  
The PermaPure drier is constructed of 2 concentric tubes. The inner tube is a special material  
that has an affinity for water vapor. The outer annulus is evacuated by the instrument pump. This  
creates a concentration gradient causing water in ambient air to diffuse into the outer annulus,  
thus the air in the inner tube becomes progressively drier as it progresses down the tube.  
Due to the large number of connections and fittings on the drier, the most common drier fault is  
leaks. Before proceding with any other procedures check the drier for leaks.  
Occasionally the drier gets contaminated. The manufacturer of the drier recommends replacing  
the drier rather than trying to clean it. If cleaning is chosen, the following options are available:  
1.  
Dirt - Clean any solids from dryer inlet by brushing. Use clean dry air to blow any loose  
particles from the inlet. Deionized water or dilute(5-10% conc HCl in deionized water)  
hydrochloric acid can be passed through the dryer. This should be done only with the dryer  
and HCl at room temperature.  
2.  
3.  
Organic liquids and freons - Rinse the inner and outer tubes using 1,1,1 trichlorethane.  
Follow the solvent with dry air to purge the solvent.  
Inorganic salts and oxgenated hydrocarbons can be removed by rinsing with 10% nitric  
acid at 50° C.  
9-53  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Figure 9-7: Ozone Generator Subsystem  
9-54  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
9.3.7 Flow/Pressure Sensor  
The flow/pressure sensor board consists of 2 pressure sensors and a flow sensor. See Figure 9-8  
for a diagram of this board. From these three sensors four values are computed and displayed on  
the front panel TEST functions. They are:  
1.  
2.  
3.  
4.  
Ozone flow - measured directly - S3, R3  
Reaction Cell Pressure - measured directly - S2, R2  
Sample gas pressure - measured directly S1, R1  
Sample Flow - computed from sample pressure and reaction cell pressure S1, S2  
The above pressures and flows are filtered to produce the front panel readings. There is a small  
delay after adjustment for a steady reading when observing the TEST functions.  
To adjust the OZONE flow:  
1.  
Go to DIAG mode by pressing SETUP-MORE-DIAG, then select SIGNAL I/O, and  
press ENTR. Select OZONE_FLOW by using NEXT-PREV keys.  
2.  
3.  
Adjust R3 so that OZONE_FLOW reads 2000 mV. This is the coarse adjustment.  
Press EXIT to return to the SAMPLE mode. Select the OZONE FL - TEST function. Use  
a calibrated flow measuring device to make small adjustments in R3 to dial in correct flow.  
To adjust the SAMPLE PRESSURE:  
1.  
Go to DIAG mode by pressing SETUP-MORE-DIAG, then select SIGNAL I/O, and  
press ENTR. Select SAMPLE_PRESS by using NEXT-PREV keys.  
2.  
3.  
Adjust R1 to read 4100 mV, which will give approximately the correct pressure.  
Press EXIT to return to the SAMPLE mode. Select SAMP - TEST function. Use a  
calibrated absolute pressure meter to make small adjustments to R1 until the correct absolute  
pressure is displayed. 29.92 in-Hg-A is the target value at sea level. The rate of decrease is  
about 1"-Hg per 1000 ft of altitude.  
To adjust the REACTION CELL PRESSURE:  
1.  
Go to DIAG mode by pressing SETUP-MORE-DIAG, then select SIGNAL I/O, and  
press ENTR. Select RCELL_PRESS by using the NEXT-PREV keys.  
2.  
3.  
Adjust R2 to read 1450 mV, which will give approximately the correct pressure.  
Press EXIT to return to the SAMPLE mode. Select RCEL - TEST function. Use a  
calibrated absolute pressure meter to make small adjustments to R2 until the correct absolute  
pressure is displayed. The target value is about 5 in-Hg-A for the external Thomas pump in  
good condition at sea level.  
9-55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
To adjust the SAMPLE FLOW  
1.  
2.  
In SAMPLE mode, scroll the TEST functions to SAMP FLW. Observe the reading. Then  
subtract the observed reading from the desired reading.  
Go to the VARS menu by SETUP-MORE-VARS, then pressing ENTR. Scroll to  
SFLOW_SET and press EDIT. The first value displayed is the SFLOW_SET:VALUE  
reading, ADD the value from step 1 to the reading shown and key in the new reading, then  
press ENTR. Check the low and high warning limits which are displayed as the next 2 values  
to make sure the new value does not exceed the warning limits. Press exit to return to  
SAMPLE mode.  
3.  
Observe the SAMP FLW value. If necessary, repeat step 2 to adjust the reading again to  
match the desired flow rate.  
9-56  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Figure 9-8: Flow/Pressure Sensor  
9-57  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Figure 9-9: NOx Sensor Module  
9-58  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Figure 9-10: NOx Sensor Module  
9-59  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
9.3.8 NOx Sensor Module  
Figure 9-9 and Figure 9-10 show the assembly of the NOx sensor module.  
9.3.8.1 PMT  
The PMT detects the light emitted by the reaction of NO with ozone. It has a gain of about  
500,000 to 1,000,000. It is not possible to test the detector outside of the instrument in the field.  
The best way to determine if the PMT is working is by using Optic Test. OT operation is  
described in Section 9.1.3.3.  
The basic method to diagnose a PMT fault is to eliminate the other components using ET, OT  
and specific tests for other sub-assemblies.  
9.3.8.2 Reaction Cell Temp  
The CPU controls the reaction cell temperature. It operates by reading a thermistor amplifier on  
the Status/Temp board. The CPU controls the temperature by toggling a bit on the V/F board.  
The V/F board TTL logic controls a solid state switch on the switch board in the PSM. The  
switched 115VAC comes out of the PSM to a connector near the underside of the reaction cell.  
A warning message may be present during initial warm-up or if the connector is not plugged in  
after cleaning the reaction cell.  
9.3.8.3 Preamp Board  
The NOx Preamp Board is a multi-function board providing circuitry to support the following  
functions.  
1.  
2.  
Preamp, buffer amplifier, physical range control hardware for the PMT detector.  
Precision voltage reference and voltage generation, and control for the PMT - HVPS  
inside the sensor module.  
3.  
4.  
5.  
Constant current generator and adjustment for the Optic Test LED.  
Voltage generation and adjustment for Electric Test.  
Thermistor amplifier, control signal generation for the PMT cooler.  
The setup and adjustment of items 1-4 above is covered in the Factory Calibration procedure in  
Section 9.1.6. Item 5 has no adjustable features.  
9-60  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Figure 9-11: PMT Cooler Subsystem  
9-61  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
9.3.8.4 PMT Cooler  
The PMT cooler uses a Peltier cooler supplied with DC current from the switching power supply  
in the Power Supply Module. An overall view is shown in Figure 9-9. A proportional  
temperature controller located on the Preamp board controls the temperature. Voltages applied to  
the cooler element vary from .1 to 12 VDC. The input voltage from the supply is 15 VDC. The  
actual -5° C setpoint will vary ± 1° C due to component tolerances. The temperature will be  
maintained within 0.1° C about the setpoint. If the temperature fails to drop after a few minutes,  
there is a problem in the cooler circuit. If the control circuit on the Preamp card is faulty a  
temperature of -10° C is reported.  
9.3.8.5 High Voltage P.S.  
The HVPS is located in the interior of the Sensor Module, and is plugged into the PMT tube. It  
requires 2 voltage inputs. The first is +15 VDC, which powers the supply. The second is the  
programming voltage, which is generated on the Preamp Board. This power supply is unlike a  
traditional PMT HVPS. It is like having 10 independent power supplies, one to each pin of the  
PMT. The test procedure below allows you to test each supply.  
Adjustment of the HVPS is covered in the Factory Calibration Procedure in Section 9.1.6. To  
troubleshoot the HVPS:  
1.  
2.  
Turn off the instrument.  
Remove the cover and disconnect the 2 connectors at the front of the NOx Sensor  
Module.  
3.  
4.  
Remove the end cap from the sensor.  
Remove the HVPS/PMT assembly from the cold block inside the sensor. Un-plug the  
PMT tube.  
5.  
6.  
7.  
Re-connect the 7 pin connector to the Sensor end cap, and power-up the instrument.  
Use Figure 9-12 to check the voltages at each pin of the supply, and the overall voltage.  
Turn off the instrument power, and re-connect the PMT tube, then re-assemble the  
sensor.  
If any faults are found in the test, you must obtain a new HVPS as there are no user serviceable  
parts inside the supply.  
9-62  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Figure 9-12: High Voltage Power Supply  
9-63  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
9.3.9 Z/S Valves  
Before troubleshooting this sub-assembly, check that the Z/S valve option was ordered, and that  
they are enabled in the software.  
Section 6.3 shows how the Z/S valves should be set-up, and how to use them with the AutoCal  
and Dynamic Cal features.  
Check for the Z/S valves:  
1.  
2.  
Check for the physical presence of the valves. See Figure 2-5 for the Z/S Valve location.  
Check front panel for option presence. The front panel should display CALS and CALZ  
buttons when the instrument is in SAMPLE mode. The presence of the buttons indicates that  
the option has been enabled in software.  
Troubleshooting the Z/S valves.  
1.  
It is possible to manually toggle each of the valves in the DIAGNOSTIC mode. Refer to  
Section 9.1.3 for information on using the DIAG mode. Also refer to Figure 8-5, Figure 8-6  
and Figure 8-7 for a pneumatic diagram of the system.  
9-64  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
9.3.10 Pneumatic System  
The pneumatic system is diagramed in Figure 8-5, Figure 8-6 and Figure 8-7.  
9.3.10.1 Leak Check  
CAUTION  
When doing a leak check do not pressurize  
the M200AU to greater than 15psi.  
Damage to internal components will  
occur at higher pressures.  
The basic cause of many performance problems is a leak. Refer to Section 8.7 for the leak check  
procedure.  
9.3.10.2 Sample Pump Diagnostic Procedures  
The sample pump is capable of maintaining the reaction cell pressure at 3-4”Hg-A. If higher  
pressures are noted, the pump may need servicing. Check the pump, inlet fittings, and analyzer  
for leaks first. If other causes have been eliminated, rebuild the pump.  
9-65  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
INTENTIONALLY BLANK  
9-66  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
10 M200AU SPARE PARTS LIST  
NOTE  
Use of replacement parts other than those supplied by Teledyne API may  
result in non-compliance with European Standard EN 61010-1.  
Table 10-1: Teledyne API M200AU Spare Parts List  
Part No.  
025690000  
000940400  
000941000  
002270100  
002730000  
002761010  
003690000  
003690100  
005140300  
007280000  
009690000  
009690100  
010680100  
010860000  
011310000  
011980000  
012360000  
013140000  
013570000  
014080100  
014610000  
016810000  
018720000  
019300000  
(table continued)  
Description  
Teledyne API Model 200AU Spare Parts List  
Orifice, 4 mil, 80 cc, Rx Cell (844-012600)  
Orifice, 13 mil 1000 cc, Rx Cell  
Gasket (Rx Cell) Qty. 12  
Window 665 NM (002-013100)  
CPU Board 200AU AMX)  
Filter, TFE, 37 mm, Qty. 100 (872-006400)  
Filter, TFE, 37 mm, Qty. 25 (872-006300)  
V/F Board  
NEW Display  
Filter, TFE, 47 mm, Qty. 100  
Filter, TFE, 47 mm, Qty. 25  
Heater, MOLY Converter  
Status/Temperature Board  
Drier Assembly Complete with Flow Control  
Assembly, MOLY Thermocouple  
Fan, Power Supply Module  
Fan, PMT Cooler  
Thermistor Assembly (Cooler)  
Assembly, High Voltage Power Supply  
Cooler Assembly  
O3 Generator Assembly W/ 01668E Driver Board  
Molybdenum converter with O3 Scrubber  
Keyboard  
10-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 10-1: Teledyne API M200AU Spare Parts List (Continued)  
Part No.  
Description  
021070000  
022220000  
FL0000001  
FL0000003  
FM0000004  
HE0000017  
HW0000020  
HW0000036  
KIT000028  
KIT000029  
KIT000029  
OR0000001  
OR0000002  
OR0000034  
OR0000042  
OR0000044  
OR0000058  
PS0000010  
025690000  
PU0000028  
PU0000030  
RL0000008  
RL0000015  
SW0000006  
SW0000008  
VA0000024  
022300000  
022930000  
023180000  
DR0000002  
PMT Pre-amplifier Board Assembly  
15 Vdc, 4A Switching Power Supply  
Sintered Filter (002-024900)  
Filter, DFU (036-040180)  
Flow Meter, 0-1000 cc  
Heater, Reaction Cell, 12W  
Spring, Flow Control  
TFE Thread Tape (48 FT)  
Retrofit , 37mm Retaining Ring, Sample Filter  
Retrofit , 47mm Retaining Ring, Sample Filter  
Retrofit , 47mm Retaining Ring, Sample Filter  
Tubing: 6’, 1/8” CLR  
O-Ring, Bearing, Cell  
O-Ring, Input/Output Mirror/Detector  
O-Ring, Sensor Assembly  
O-Ring, Reaction Cell  
O-Ring, Sample Filter  
15V Switching Power Supply  
Teledyne API Model 200AU Spare Parts List  
Pump, 115V/60Hz, M200AU  
Pump Rebuild Kit, M200AU  
Solid State Relay, 12 Vdc  
Solid State Relay, 115 Vac  
Overheat SW, Cell/Oven  
Pressure Sensor  
Manifold Valve, 3-Way, Vent  
DC Power Supply Board  
Instruction Manual for M200AU  
M200AU Expendables Kit  
PMT Desciccant Baggies  
10-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table 10-2: Teledyne API MODEL 200AU Expendable Kit  
Part No.  
Description  
023180000  
Includes:  
M200AU Expendables Kit  
Qty  
4
002270100  
003690100  
FL0000001  
FL0000003  
HW0000020  
OR0000001  
Gasket (Rx Cell) Qty. 12  
Filter, TFE, 37 mm, Qty. 25 (872-006300)  
Sintered Filter (002-024900)  
Filter, DFU (036-040180)  
Spring, Flow Control  
1
2
1
2
O-Ring, Flow Control  
4
10-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
INTENTIONALLY BLANK  
10-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
APPENDIX A ELECTRICAL SCHEMATICS  
Table A-1: Electrical Schematics  
Part No.  
00402  
00403  
00514  
00515  
00704  
00705  
01086  
01087  
01109  
01110  
01114  
01115  
0113912  
01206  
01207  
01464  
01465  
01471  
01668  
01669  
01839  
01840  
01916  
01917  
01930  
Name  
Sensor Board Assembly  
Sensor Board Schematic  
V/F Board Assembly  
V/F Board Schematic  
Keyboard Assembly  
Keyboard Schematic  
Status/Temp Assembly  
Status/Temp Schematic  
Motherboard Assembly  
Motherboard Schematic  
Connector Board Assembly  
Connector Board Schematic  
PSM Overall schematic - CE Mark  
Switch Board Assembly  
Switch Board Schematic  
DC Power Supply Assembly  
DC Power Supply Schematic  
4-20 mA Output Option  
Ozone Generator Power Supply Assembly  
Ozone Generator Power Supply Schematic  
Thermoelectric Cooler Control Assembly  
Thermoelectric Cooler Control Schematic  
Connector Board Schematic - CE MARK  
Connector Board Assembly - CE MARK  
Keyboard Assembly - CE MARK  
Keyboard Schematic - CE MARK  
01931  
(table continued)  
A-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Teledyne API Model 200AU NOX Analyzer Instruction Manual, 02293, Rev. F  
Table A-1: Electrical Schematics (Continued)  
Part No.  
02034  
02035  
02037  
02038  
02107  
02108  
02222  
02223  
02230  
02231  
03172  
03173  
03174  
03175  
Name  
A-Series Valve Driver Assembly  
A-Series Valve Driver Schematic  
ISBX I/O Port Assembly  
ISBX I/O Port Schematic  
Preamp Board Assembly  
Preamp Board Schematic  
Switch Board Assembly - CE MARK  
Switch Board Schematic - CE MARK  
DC Power Supply Assembly - CE MARK  
DC Power Supply Schematic - CE MARK  
M200A Interconnect Diagram  
M200AH Interconnect Diagram  
M200AU Interconnect Diagram  
M201A Interconnect Diagram  
A-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Toastmaster Toaster T75B User Manual
Toshiba Cable Box 58L9300UC User Manual
Toshiba CRT Television 27A40 User Manual
Toshiba Flat Panel Television 23L1350U User Manual
Tripp Lite TV Cables U021 003 User Manual
Uniden Cordless Telephone D17882 User Manual
Uniden Cordless Telephone DECT4096 User Manual
ViewSonic Flat Panel Television VS12121 1M User Manual
Vizio TV Mount XMF1000 User Manual
VTech Cordless Telephone 20 User Manual