INSTRUCTION MANUAL
MODEL 460L
Nema Ozone Monitor
9480 CARROLL PARK DRIVE
SAN DIEGO, CA 92121-5201
USA
Toll-free Phone: 800-324-5190
Phone: 858-657-9800
Fax: 858-657-9816
Email: [email protected]
05012 REV. D
DCN 5164
© 2008 T-API
22 January 2009
PRINT DATE: 19 February 2009
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
460L Instruction Manual
Safety Messages
SAFETY MESSAGES
Your safety and the safety of others is very important. We have provided many important safety
messages in this manual. Please read these messages carefully.
A safety message alerts you to potential hazards that could hurt you or others. Each safety
message is associated with a safety alert symbol. These symbols may be found in the manual and
inside the monitor. The definition of these symbols is described below:
GENERAL SAFETY HAZARD: Refer to the instructions for details on the
specific hazard.
CAUTION: Hot Surface Warning
CAUTION: Electrical Shock Hazard
TECHNICIAN SYMBOL: All operations marked with this symbol are to
be performed by qualified maintenance personnel only.
CAUTION
The monitor should only be used for the purpose and in the manner described in this
manual. If you use the monitor in a manner other than that for which it was intended,
unpredictable behavior could ensue with possible hazardous consequences.
NOTE
Technical Assistance regarding the use and maintenance of the Model 460L Nema Ozone
Monitor or any other Teledyne Instruments product can be obtained by:
Contacting Teledyne Instruments’ Customer Service Department at 800-324-5190
or
Via the internet at http://www.teledyne-api.com/inquiries.asp
05228 Rev B
DCN 5164
i
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
Safety Messages
460L Instruction Manual
THIS PAGE IS INTENTIONALLY LEFT BLANK
ii
05228 Rev B
PRINTED DOCUMENTS ARE UNCONTROLLED
DCN 5164
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
460L Instruction Manual
Table of Contents
TABLE OF CONTENTS
1. INTRODUCTION........................................................................................................................... 1
3.1. Unpacking..............................................................................................................................7
3.2. Mechanical Installation.............................................................................................................9
3.3. Pneumatic Connections ..........................................................................................................10
3.4. Electrical Connections ............................................................................................................12
3.4.1. AC Power Connection.......................................................................................................12
3.5. Signal I/O Connections...........................................................................................................13
3.5.1. Analog Output ................................................................................................................13
3.5.2. Digital Status Outputs......................................................................................................14
3.5.3. Control Inputs ................................................................................................................15
3.5.4. Relay Outputs.................................................................................................................15
3.6. Initial Startup .......................................................................................................................17
3.7. Setting up the Serial Communications Port................................................................................17
3.7.1. Physical Serial Port Configuration ......................................................................................18
3.7.2. Software Setup for Serial Port Communications...................................................................19
5.1. Sample Conditioning System...................................................................................................23
5.2. Current Loop Analog Output....................................................................................................23
5.3. Ozone Destruct Option...........................................................................................................23
6.1. Front Panel Display................................................................................................................25
6.1.1. O readout ....................................................................................................................25
6.1.2. Zero Point Calibration......................................................................................................25
6.1.3. Status LED’s...................................................................................................................25
6.2. Concentration Alarms.............................................................................................................26
6.2.1.1. Concentration Alarm Configuration..............................................................................26
6.2.1.2. Remotely Sensing the Status of the Alarm....................................................................26
6.2.1.3. Clearing Alarms........................................................................................................26
7.1. Serial Port Command Syntax...................................................................................................27
7.2. Serial Port Command Summary...............................................................................................28
7.3. Serial Port Command Reference ..............................................................................................29
7.3.1. ALMACK.........................................................................................................................29
7.3.2. ALMSTAT.......................................................................................................................30
7.3.3. CZERO ..........................................................................................................................31
7.3.4. DACSTEP.......................................................................................................................32
7.3.5. O3.................................................................................................................................33
7.3.6. SETADDR.......................................................................................................................34
7.3.7. TDUMP ..........................................................................................................................35
7.3.8. TLIST............................................................................................................................36
7.3.9. VGET ............................................................................................................................37
7.3.10. VLIST..........................................................................................................................38
7.3.11. VSET...........................................................................................................................39
8. CALIBRATION.............................................................................................................................41
8.1. Zero Point calibration.............................................................................................................41
8.2. Span Point Calibration............................................................................................................41
8.3. Adjusting the Optional Current Loop Output ..............................................................................43
9.1. Replacing the Particulate Filter Element ....................................................................................45
05228 Rev B
DCN 5164
iii
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
Table of Contents
460L Instruction Manual
9.1.1. Filter Replacement Procedure............................................................................................45
9.1.2. Mounting the Particulate Filter Externally............................................................................46
9.2. Maintaining the Optional H2O Coalescing Filter...........................................................................47
9.2.1. Draining the Optional Coalescing Filter ...............................................................................47
9.2.2. Replacing the Coalescing Membrane ..................................................................................48
9.3. UV Lamp Replacement ...........................................................................................................49
9.4. Cleaning Exterior Surfaces of the 460L .....................................................................................50
9.5. Degree of Protection..............................................................................................................50
10.1. Basic O3 Measurement Principle .............................................................................................51
10.1.1. (Beer’s Law).................................................................................................................51
10.1.2. The Absorption Path ......................................................................................................52
10.1.3. The Reference / Measurement Cycle ................................................................................53
10.1.4. Digital Noise Filter.........................................................................................................54
10.2. Pneumatic Theory of Operation..............................................................................................55
10.2.1. Basic Pneumatic Flow And Flow Control ............................................................................55
10.2.2. Internal Pump and Flow Control.......................................................................................55
10.2.3. Optional Sample Conditioning .........................................................................................56
10.2.3.1. H2O Coalescing Filter...............................................................................................56
10.2.3.2. H2O Vapor Dryer.....................................................................................................56
10.2.3.3. Ozone Destruct Scrubber .........................................................................................57
10.3. Electronic Theory of Operation...............................................................................................58
10.3.1. Main Board...................................................................................................................58
10.3.2. O3 Sensor Module..........................................................................................................59
10.3.2.1. O3 Sensor Components............................................................................................59
10.3.2.2. Sensor Module PCA .................................................................................................60
10.3.3. CPU Board....................................................................................................................61
10.3.3.1. I/O Functions .........................................................................................................61
10.3.3.2. Status and Alarm Functions......................................................................................61
10.3.4. Display Driver and Keyboard Assembly.............................................................................62
10.3.4.1. Keyboard...............................................................................................................62
10.3.4.2. Display..................................................................................................................62
10.3.4.3. Display Driver ........................................................................................................62
10.3.5. Power Distribution.........................................................................................................63
11.1. Status Output Summary.......................................................................................................65
11.2. Troubleshooting Using Status Outputs ....................................................................................65
11.2.1. Sensor OK....................................................................................................................65
11.2.2. Invalid Reading.............................................................................................................66
11.2.3. Lamp Low ....................................................................................................................66
11.2.4. Status LED / Status Output Troubleshooting Summary .......................................................66
11.3. Concentration Alarm Outputs.................................................................................................67
11.4. Technical Assistance ............................................................................................................68
12.1. How Static Charges are Created.............................................................................................69
12.2. How Electro-Static Charges Cause Damage .............................................................................70
12.3. Common Myths About ESD Damage .......................................................................................71
12.4. Basic Principles of Static Control ............................................................................................72
12.4.1. General Rules ...............................................................................................................72
12.4.2. Basic anti-ESD Procedures for Analyzer Repair and Maintenance ..........................................73
12.4.2.1. Working at the Instrument Rack................................................................................73
12.4.2.2. Working at an Anti-ESD Work Bench..........................................................................73
12.4.2.3. Transferring Components from Rack to Bench and Back................................................74
12.4.2.4. Opening Shipments from Teledyne Instruments’ Customer Service.................................75
12.4.2.5. Packing Components for Return to Teledyne Instruments Customer Service ...................76
iv
05228 Rev B
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
460L Instruction Manual
Table of Contents
List of Figures
M460L Front Cover Layout.............................................................................7
460L Layout ................................................................................................8
460L Sensor Module Layout ...........................................................................9
Alarm Output Relays................................................................................... 16
O3 Absorption Path ..................................................................................... 53
LIST OF TABLES
460L Specifications.......................................................................................5
460L Ventilation Clearances......................................................................... 10
Relay Output Operation............................................................................... 16
Status LED’s.............................................................................................. 25
Measurement / Reference Cycle.................................................................... 54
05228 Rev B
DCN 5164
v
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
Table of Contents
460L Instruction Manual
LIST OF APPENDICES
Appendix A – Spare Parts List
Appendix B – List of Schematics
vi
05228 Rev B
PRINTED DOCUMENTS ARE UNCONTROLLED
DCN 5164
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
460L Instruction Manual
Introduction
1. INTRODUCTION
1.1. Preface
Teledyne Instruments is pleased that you have purchased the Model 460L NEMA Ozone Monitor.
Included is a full one-year warranty (see Section 2.2) and we at Teledyne Instruments will be
pleased to provide you with any support required so that you may utilize our equipment to the
fullest extent.
The Model 460L is a microprocessor based low range ozone monitor for safety monitoring of ozone
levels in a variety of applications such as water treatment, food processing, and research. The
design has been specifically optimized for applications requiring the measurement of ozone at the
typically low concentration levels encountered when tracking ambient conditions. The Model 460L
has been designed to give accurate and stable readings over long time periods with little or no
maintenance or calibration.
The flexibility of the software as well as the analog and digital I/O allow the Model 460L to
interface with a broad range of devices for process control and data logging.
We hope you will not experience any problems with the Teledyne Instruments Model 460L but if
you do, our full time customer service department is always available to answer your questions.
1.2. 460L Documentation
The documentation for this monitor is available in several different formats:
Printed format, part number 050120100
Electronic format on a CD-ROM, part number 050120200
The electronic manual is in Adobe® Systems Inc. “Portable Document Format”. The Adobe®
Acrobat Reader® software, which is necessary to view these files, can be downloaded for free from
the Internet at http://www.adobe.com/.
The electronic version of the manual has many advantages:
Keyword and phrase search feature
Figures, tables and internet addresses are linked so that clicking on the item will display
the associated feature or open the website.
A list of chapters and sections as well as thumbnails of each page are displayed to the left
of the text.
Entries in the table of contents are linked to the corresponding locations in the manual.
Ability to print sections (or all) of the manual
05228 Rev B
DCN 5164
1
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
Introduction
460L Instruction Manual
1.3. Using This Manual
This manual has the following data structures:
1.0 Table of Contents:
Outlines the contents of the manual in the order the information is presented. This is a good
overview of the topics covered in the manual. There is also a list of tables, a list of figures and a
list of appendices. In the electronic version of the manual, clicking on any of these table entries
automatically views that section.
2.0 Specifications and Warranty
This section contains a list of the monitor’s performance specifications, a description of the
conditions and configuration under which Teledyne Instruments Incorporated warranty statement
applies.
3.0 Getting Started:
A concise set of instructions for setting up, installing and starting your monitor for the first time.
This includes unpacking; mechanical installation; attaching all pneumatic lines; attaching all
electrical and electronic connections and the physical configuration the RS-232/RS-485 port.
4.0 FAQ:
Answers to the most frequently asked questions about operating the monitor.
5.0 Optional Hardware & Software
A description of optional equipment to add functionality to your monitor.
6.0 Operation Instructions
Instructions for operating the monitor and using its basic features and functions.
7.0 Serial Communications
Information regarding the syntax and command definitions for the monitor’s serial I/O interface.
8.0 Calibration Procedures
General information and step-by-step instructions for manually calibrating your monitor.
9.0 Monitor Maintenance
Description of certain preventative maintenance procedures that should be regularly performed on
your monitor to keep it in good operating condition.
10.0 Theory of Operation
An in-depth look at the various principals by which your monitor operates as well as a description
of how the various electronic, mechanical and pneumatic components of the monitor work and
2
05228 Rev B
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
460L Instruction Manual
Introduction
interact with each other. A close reading of this section is invaluable for understanding the
monitor’s operation.
11.0 Troubleshooting Section:
This section includes pointers and instructions for diagnosing problems with the monitor, such as
excessive noise or drift, as well as instructions on performing repairs of the monitor’s major
subsystems.
12.0 Electro-Static Discharge (ESD) Primer:
This section describes how static electricity occurs; why it is a significant concern and how to
avoid it and; how to avoid allowing ESD to affect the reliable and accurate operation of your
monitor.
USER NOTES:
05228 Rev B
3
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
Introduction
460L Instruction Manual
USER NOTES:
4
05228 Rev B
PRINTED DOCUMENTS ARE UNCONTROLLED
DCN 5164
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
460L Instruction Manual
Specifications and Warranty
2. SPECIFICATIONS AND WARRANTY
2.1. Specifications
Table 2-1 460L Specifications
Measurement Principle
Ranges
UV Absorption (Beer Lambert Law)
1ppm to 500 ppm: User selectable
ppm; ppb
Measurement Units
Accuracy
1% of Full Scale
Zero Noise
< 0.0015 ppm (rms)
Span Noise
< 0.5% of reading (rms) (above 0.1 ppm)
< 0.003 ppm (rms)
Lower Detectable Limit
Linearity
Better than 1% of selected range
<30 sec
Response Time (95%)
Repeatability
0.5% of selected range
0001 ppm, 1 ppb
Display Resolution
Gas Flow Rate
1.0-2.0 LPM
Compensation
Pressure, Temperature (NTP = 273.15K, 760mmHg)
11.0 – 16.0 psia
Gas Inlet Pressure Range
Temperature Range
Humidity Range
Dimensions (H x W x D)
5-45 ˚C
10-90% RH, Non-Condensing
12.64" x 11.19" x 6.08"
(321mm x 284mm x 154mm)
Weight
9.40lb (4.27kg)
Power
110-240V~, 50/60Hz, 200 VA
Environmental Conditions
Installation Category (Over Voltage Category) II
Pollution Degree 2
Maximum Operating Altitude
Analog Voltage Output
2000 meters
Single output: 0-5V,
Option available converting voltage output to 4-20mA output
with maximum voltage between outputs and ground 60V peak
Relay Outputs
3 relay outputs: Sensor OK and two concentration alarms (HI &
Hi-Hi)
Relay type & Output Rating
SPDT: 250V AC, 3A
IP65 (NEMA 4X)
Degree of Protection (IP Code)
05228 Rev B
DCN 5164
5
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
Specifications and Warranty
460L Instruction Manual
2.2. Warranty
WARRANTY POLICY (02024D)
Prior to shipment, T-API equipment is thoroughly inspected and tested. Should equipment failure
occur, T-API assures its customers that prompt service and support will be available.
COVERAGE
After the warranty period and throughout the equipment lifetime, T-API stands ready to provide
on-site or in-plant service at reasonable rates similar to those of other manufacturers in the
industry. All maintenance and the first level of field troubleshooting is to be performed by the
customer.
NON-API MANUFACTURED EQUIPMENT
Equipment provided but not manufactured by T-API is warranted and will be repaired to the
extent and according to the current terms and conditions of the respective equipment
manufacturers warranty.
GENERAL
During the warranty period, T-API warrants each Product manufactured by T-API to be free from
defects in material and workmanship under normal use and service. Expendable parts are
excluded.
If a Product fails to conform to its specifications within the warranty period, API shall correct such
defect by, in API's discretion, repairing or replacing such defective Product or refunding the
purchase price of such Product.
The warranties set forth in this section shall be of no force or effect with respect to any Product:
(i) that has been altered or subjected to misuse, negligence or accident, or (ii) that has been used
in any manner other than in accordance with the instruction provided by T-API, or (iii) not
properly maintained.
THE WARRANTIES SET FORTH IN THIS SECTION AND THE REMEDIES THEREFORE ARE
EXCLUSIVE AND IN LIEU OF ANY IMPLIED WARRANTIES OF MERCHANTABILITY, FITNESS FOR
PARTICULAR PURPOSE OR OTHER WARRANTY OF QUALITY, WHETHER EXPRESSED OR IMPLIED.
THE REMEDIES SET FORTH IN THIS SECTION ARE THE EXCLUSIVE REMEDIES FOR BREACH OF
ANY WARRANTY CONTAINED HEREIN. API SHALL NOT BE LIABLE FOR ANY INCIDENTAL OR
CONSEQUENTIAL DAMAGES ARISING OUT OF OR RELATED TO THIS AGREEMENT OF T-API'S
PERFORMANCE HEREUNDER, WHETHER FOR BREACH OF WARRANTY OR OTHERWISE
6
05228 Rev B
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
460L Instruction Manual
Getting Started
3. GETTING STARTED
3.1. Unpacking
1. Inspect the received packages for external shipping damage. If damaged, please advise
the shipper first, then Teledyne Instruments.
2. Loosen the 2 setscrews located in the top and bottom left corners of the front and swing
open the cover.
Set Screw
O3 Concentration
Display
SYSTEM OK
Status
INVALID READING
LAMP LOW
LED’s
ALARM ACTIVE
ZERO
Buttons
Ozone Monitor – Model 460L
Alarm
Acknowledge
Key
Gas
Flowmeter
&
Control
Set Screw
Figure 3-1 M460L Front Cover Layout
3. Inspect the interior of the monitor to make sure all circuit boards and other components
are in good shape and properly seated.
NOTE
Printed circuit assemblies (PCAs) are static sensitive. Electro-static discharges (ESD),
too small to be felt by the human nervous system, are large enough to destroy sensitive
circuits.
Before touching PCAs, read Chapter 12 of this manual and follow the procedure
described there for avoiding damage to your monitor due to ESD.
05228 Rev B
DCN 5164
7
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
Getting Started
460L Instruction Manual
CAUTION
Never disconnect electronic circuit boards, wiring harnesses or
electronic subassemblies while the unit is under power.
4. Check the connectors of the various internal wiring harnesses and pneumatic hoses to
make sure they are firmly and properly seated (see Figure 3-2).
5. Verify that all of the optional hardware ordered with the unit has been installed. These are
listed on the paperwork accompanying the monitor.
CPU
Board
Analog Relay H2O Vapor Dryer
(optional)
Outputs
Optional
Ozone
Scrubber
Not Shown
Signal I/O
Connector
Main
Board
AC Power
Input
Connector
Display
Board
Flow Meter
Sensor Module
H2O
Coalescing
Filter
(optional)
Main
Power
Supply
Particulate
Filter
Pump
Ozone
Inlet
Exhaust
Outlet
Figure 3-2 460L Layout
8
05228 Rev B
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
460L Instruction Manual
Getting Started
UV
Lamp
UV Lamp
Housing
UV Lamp
Power Connector
Ozone Outlet
Manifold
UV Lamp
Power Supply
Reference
Scrubber
Absorption
Tube
Sensor
Module PCA
Ozone Intlet
Manifold
Measure / Reference
Valve
UV
Detector
Figure 3-3 460L Sensor Module Layout
3.2. Mechanical Installation
Mount the enclosure securely to a vertical surface.
Figure 3-4 below shows the locations of the four mounting holes.
All four mounting holes should be used to secure the monitor.
Use stainless steel, 5/16” diameter bolts.
VENTILATION CLEARANCE:
When installing the monitor be sure to leave sufficient ventilation clearance.
05228 Rev B
9
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
Getting Started
460L Instruction Manual
Table 3-1
460L Ventilation Clearances
AREA
MINIMUM REQUIRED CLEARANCE
1 inch
1 inch
Sides of the monitor
Above and below the monitor.
Figure 3-4 460L Mounting Hole Locations and Dimensions
3.3. Pneumatic Connections
1. Connect a ¼” exhaust line to the fitting labeled ‘Exhaust.’
CAUTION
If using the 460L to measure O3 levels that are ≥100 ppb exhaust gas
from the MODEL 460L may contain dangerous levels of ozone!
The optional ozone destruct (see Section 5.3) should be installed
and
Make sure that the exhaust line is vented to an outside area.
2. Connect the ozone delivery line to the ¼” inlet fitting labeled “Ozone Inlet” on the bottom
face of the enclosure (See Figure 3-5.)
10
05228 Rev B
PRINTED DOCUMENTS ARE UNCONTROLLED
DCN 5164
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
460L Instruction Manual
Getting Started
NOTE
The ozone delivery pressure should be at ambient pressure +/- 5 PSIG.
All tubing used should be made of ozone resistant material such as Stainless Steel, PTFE
(Teflon) or FEP. API can supply appropriate tubing for connecting the ozone supply
line.
Display
Ozone
Scrubber
(optional)
Gas
Flowmeter
&
Coalescence
Control
Filter
(optional)
OZONE INLET
EXHAUST
OUTLET
Install AC
Power Cord
Through Here
Figure 3-5 460L Pneumatic Connections
The gas flow rate through the monitor should be established between 1.0 and 2.0 L/min.
05228 Rev B
DCN 5164
11
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
Getting Started
460L Instruction Manual
3.4. Electrical Connections
NOTE
It is recommended that if multi-strand wires are used for the following electrical
connections. To ensure a reliable connection the wires should be:
“Tinned” with solder or;
Terminated with insulated crimped ferrules, such as Entrelec® P/N 304.456.02
(18 gauge) or 304.558.10 (22 gauge)
3.4.1. AC Power Connection
CAUTION
Disconnect power to the AC mains before making or removing any
electrical connections to the 460L.
CAUTION
A proper earth ground connection must be made to the receptacle
labeled “Earth Ground” on the 3-pin AC connector. Failure to do so may
result in a shock hazard and malfunction of the monitor
Connect AC power to the monitor.
3. Install a ½” conduit fitting for routing the electrical wiring into the monitor through the
hole provided (see Figure 3-5). In order to maintain the IP (NEMA4X) rating of the
enclosure, an appropriate sealed conduit connector should be used.
4. Attach the leads of the power line to the AC power connector (see Figure 3-6)
Analog and
Digital I/O
Connector
Relay
Outputs &
Connectors
Ground
AC Line
AC Neutral
AC POWER
Connector
Figure 3-6 Location of Electrical Connectors
12
05228 Rev B
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
460L Instruction Manual
Getting Started
3.5. Signal I/O Connections
All digital and analog signal I/O connections are made via a 16-pin connector on the main board
(See Figure 3-6 for the location of the connector.)
This connector can be unplugged from the header on the main board for easier access when
wiring. To disconnect from main board, loosen the two retaining screws at either end of the
connector.
Retaining Screw
Analog Out +
Analog Out -
Zero Input
Aux Input
Gnd
Status Out 1
Status Out 2
Status Out 3
Status Out 4
Status Com
Serial TX
Serial RX
Serial GND
Status Out 5
Status Out 6
Retaining Screw
Figure 3-7 Signal I/O Connector Pin Assignments
3.5.1. Analog Output
The 460L is equipped with one analog output that is factory configurable as either a 0-5 VDC
signal or a 4-20 mA signal. You may verify how your 460L is set up by checking the information
on the monitor’s serial number tag.
The analog output requires two connections: ANALOG +, the signal line, and ANALOG–, the
ground connection. See Figure 3-7 for the locations of the ANALOG OUT + and ANALOG OUT–
pins.
05228 Rev B
DCN 5164
13
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
Getting Started
460L Instruction Manual
3.5.2. Digital Status Outputs
The 460L has six assigned digital status outputs for indicating error and operational status
conditions of the monitor as well as the status of its O3 concentration alarms (see Table 3-2).
These outputs are in the form of opto-isolated open-collector transistors. They can be used to
drive status LED’s on a display panel or interface to a digital device such as a Programmable Logic
Controller (PLC). Several of the status outputs are useful tools for diagnosing sensor and system
level malfunctions (see Section 11.2 for more information).
Table 3-2 Digital Status Output Descriptions
LABEL1
NAME
Sensor O.K.
OPERATION
Normally On
Normally Off
Normally Off
Normally Off
Normally Off
Normally Off
N/A
STATUS OUT 12
STATUS OUT 22
STATUS OUT 32
STATUS OUT 4
STATUS OUT 5
STATUS OUT 6
STATUS COM
Invalid Reading
Lamp Low
Alarm Active
HI Alarm Status
HI-HI Alarm Status
Common Pin for all Status Outputs
1
2
See Figure 3-7 for pin locations of the these output lines on the monitor’s 16-pin I/O
connector
See Section 11-2 for definitions and interpretations of these output.
Figure 3-8 shows the most common way of connecting the digital outputs to an external device
such as PLC.
460M
PLC OR OTHER DEVICE
+5V
Status Outputs 1-6 (Collector)
STATUS COM (Emitter)
Digital Input
Opto-Isolator
Figure 3-8 Digital Status Output Connections
Note
Most devices, such as PLC’s, have internal provision for limiting the current that the
input will draw from an external device.
When connecting to a unit that does not have this feature, external dropping-resistors
must be used to limit the current through the transistor output to 50mA or less.
At 50 mA, the transistor will drop approximately 0.2V from its collector to emitter.
14
05228 Rev B
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
460L Instruction Manual
Getting Started
3.5.3. Control Inputs
Two digital control inputs are also available for use on the 460L. The control inputs are used for
remote control of the 460L by a device such as a PLC. They are labeled ZERO INPUT and AUX
Table 3-3 Control Inputs
INPUT
DESCRIPTION
ZERO INPUT
AUX INPUT
This input performs exactly the same function as the ‘Zero’ buttons on the front panel.
This input can be used to clear active concentration alarms in similar fashion as the serial
communications ALMACK.
Signal I/O
Connector
Analog Out +
Analog Out -
Zero Input
Aux Input (AlmAck)
Gnd
Figure 3-9 Control Input Connections
NOTE:
Never connect a voltage level output from another device to these contacts.
3.5.4. Relay Outputs
The 460L is equipped with three SPDT relays. They are located at the top right hand side of the
main board and are labeled RELAY 1, RELAY 2 & RELAY 3.
RELAY 1 corresponds to the Sensor OK status output and LED;
RELAY 2 corresponds to the HI concentration alarm, and;
RELAY 3 corresponds to the HI-HI concentration alarm
See Section 6.2 for more information on these alarms.
05228 Rev B
15
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
Getting Started
460L Instruction Manual
Below each relay is a 3-pin connector that allows the relay to be connected for either
normally open or normally closed operation.
Table 3-4 describes how to connect the alarm relays.
NOTE
The relay contacts are rated to 3A at 240VAC. Do not exceed these ratings when
connecting equipment to the instrument.
RELAY 1
RELAY 2
RELAY 3
D6
D7
D8
STATUS LEDS
Figure 3-10 Alarm Output Relays
Table 3-4 Relay Output Operation
RELAY PIN
STATE2
STATUS
RELAY
FUNCTION
COMMENTS
LED1
C
O
M
N.
O.
N.
C.
D6 ON
460L is operating normally
SENSOR OK ON
1
Problem with the O3 sensor module.
See Section 11.2.
D6 OFF
SENSOR OK OFF
D7 ON
O3 concentration > HI alarm limit
O3 concentration < HI alarm limit
HI Alarm ON
2
3
D7 OFF
HI Alarm OFF
D8 ON
O3 concentration > HI-HI alarm limit
O3 concentration < HI-HI alarm limit
HI-HI Alarm ON
D8 OFF
HI-HI Alarm OFF
16
05228 Rev B
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
460L Instruction Manual
Getting Started
1
Located just below each relay connector (see Figure 3-10)
N.O. = Normally Open operation.
2
N.C. = Normally Closed operation.
3.6. Initial Startup
Perform the following steps when initially starting up the 460L or when bringing the monitor back
into service it has been shut down for repair or maintenance.
1. Turn on power to the monitor.
The display will briefly display the “API” logo followed by the software version.
The display will then begin showing ozone concentration.
2. Establish a flow of ozone to the monitor.
Flow rate through the monitor should be between 0.5 – 2.0 LPM (Liters per minute.)
Adjust as needed using the needle valve of the flow meter on the front panel
(see Figure 3-1).
3. Let the monitor warm up for a minimum of 5 minutes.
4. Check Status LED’s on front panel;
Sensor OK LED should be ON.
All other LED’s should be OFF.
If the Status LED’s are not in this state, refer to Chapter 11 for troubleshooting
information.
5. Observe the monitor for several more minutes at zero to ensure that it is stable.
THE MONITOR IS NOW READY TO MEASURE OZONE.
3.7. Setting up the Serial Communications Port
The 460L’s bi-directional RS-232/485 Serial Port Interface allows the user to communicate with
the monitor via a computer over that computer’s serial communications port (COM port). A
terminal emulation program such as HyperTerminal is required to be installed and running on the
host computer.
The following three pins are provided on the I/O connector for serial communications (see Figure
Table 3-5 Serial I/O Port Connection
LABEL
DESCRIPTION
Serial Transmit (RS-485 – A)
Serial Receive (RS-485 – B)
Serial Ground (RS-232 Only)
SERIAL TX
SERIAL RX
SERIAL GND
05228 Rev B
DCN 5164
17
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
Getting Started
460L Instruction Manual
While the standard factory configuration is for RS-232, the monitor’s serial port can be configured
Use RS-232 for direct connection to a nearby (no more than 6-8 feet cable length) PC or
Laptop, RS-232 should be used.
Use RS-485 for permanent connection to continuously operating data acquisition systems
or connections over greater distances.
3.7.1. Physical Serial Port Configuration
6. To configure the com port for RS-232 or RS-485, move the 4 shunts on JP3 of the CPU PCA
(P/N #03492) to the proper position as shown in below.
The jumpers may already be in this position but this still needs to be verified.
Also make sure that JP1 is jumpered. It may be hanging off of one pin, make sure it is
JP2 can either be jumpered or not as it is already shorted on the board.
Reset Button
Micro-Controller
Connector for
Optional 4-20 mA
output
JP2
JP1
JP3 set for RS-232
JP3 set for RS-485
Figure 3-11 RS-232/RS-485 Jumper Location and Settings
18
05228 Rev B
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
460L Instruction Manual
Getting Started
7. Connect the appropriate type of cable to the 16 pin Signal I/O connector inside the
analyzer.
construct wiring for both RS-232 and RS-485 connections.
DB9 Female
RS-232
1
6
Serial TXD
2
7
Serial RXD
3
8
4
9
Serial GND
5
RS-485
Twisted Pair
Serial TXD
Serial RXD
RS-485-A
RS-485-B
Figure 3-12 Typical RS-232 and RS-485 Connections
3.7.2. Software Setup for Serial Port Communications
NOTE
This section refers to various serial communication commands for the M460M. For
detailed information regarding these commands see Chapter 7.
1. Connect the other end of the cable to your serial Com port on your computer
2. Open up Hyper-terminal or another terminal program and set up a connection with the
settings below. Your com port is most likely COM1.
05228 Rev B
19
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
Getting Started
460L Instruction Manual
The Serial Port of the device being used to communicate with the 460L should be
configured as follows:
Table 3-6 Serial Port Configuration
PARAMETER
Baud Rate
Data Bits
Stop Bits
Parity
VALUE
9600 bps
8
1
None
3. Find the address of the 460L.
possible address (0-9) until the M460M responds:
EXAMPLE: 0O3<CR>
1O3<CR>
2O3<CR>
…
9O3<<CR>
When the instrument responds record the first number in the response line
This is the address of the analyzer.
4. To determine if the M460L is operating correctly type:
1TLIST<CR>
This analyzer will respond with a list of the current values of its test functions. See Section
7.3.8 for a list of the nominal values for these functions
5. Check the analyzer’s operation variables (VARS). These describe certain application
specific conditions such as units of measure, measurement ranges, etc.
To check the current state of the VARS, type:
1VLIST<CR>
This analyzer will respond with a list of the current values of its VARS. See Section 7.3.10
for a list of the nominal values for these functions
6. If you need to change one of the VARS settings, see the VSET command (see Table 7-3 of
7. After viewing the data you can return the analyzer to normal operation.
After changing the value of any of the VARS return the analyzer to normal operation.
The simplest way to do this is to turn the analyzer off and back on. User Notes:
USER NOTES:
20
05228 Rev B
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
460L Instruction Manual
Frequently Asked Questions
4. FREQUENTLY ASKED QUESTIONS
Q: What do I do if I smell Ozone and suspect a leak?
A: Ensure that all the fittings are tight. If the fittings are tight and ozone is still detected send
the monitor back for repair to Teledyne Instruments’ customer service for repair.
Q: What do I do if my CPU status light stops flashing?
A: The CPU has stopped working. This is a major malfunction of the monitor. Return it to
Teledyne Instruments’ customer service for repair.
Q: What do I do if the Status OK light turns off or doesn't turn on after 30 min.?
A: If the status ok light is off and the Lamp low light is on then the monitor’s UV lamp needs to
be adjusted.
If the status ok light is off and the Lamp Low light is off then most likely the UV lamp output
has drifted to >5000mv and needs to be adjusted.
Q: What does it mean if the Invalid Reading light turns on?
A: This will happen if the:
Ozone supply pressure exceeds 14.9 psia.
Ozone concentration exceeds the Range of the monitor then this light will turn on.
Ozone concentration goes excessively negative.
Q: What do I do if my Lamp Low light turns on?
A: If the Lamp Low light turns on and the Status OK light is ON, the UV reference value has
dropped to < 1000mv. The monitor will continue to run with no problems until the UV
reference drops below 500mv.
If the Lamp low light turns on and the Status OK light is OFF, the UV reference has dropped
below acceptable limits and will have to be adjusted.
Q: When should I change the Particulate Filter and how do I change it?
A: The Particulate filter should be changed monthly. See Section 9.1 for instructions on
performing this replacement.
05228 Rev B
21
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
Frequently Asked Questions
460L Instruction Manual
USER NOTES:
22
05228 Rev B
PRINTED DOCUMENTS ARE UNCONTROLLED
DCN 5164
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
Optional Hardware and Software
460L Instruction Manual
5. OPTIONAL HARDWARE AND SOFTWARE
This section includes descriptions of the hardware and software options available for the 460L
monitor. For assistance with ordering these options please contact the sales department of
Teledyne Instruments at:
TOLL-FREE: 800-324-5190
TEL: +1 858-657-9800
FAX: +1 858-657-9816
E-MAIL: [email protected]
5.1. Sample Conditioning System
This option is required for 460L’s that will be used in applications where the sample gas includes
liquid or vaporous water. This option includes two major components:
A coalescing water drop-out filter, and;
A permeation tube dryer.
5.2. Current Loop Analog Output
This option adds isolated, voltage-to-current conversion circuitry to the monitor’s CPU card. This
option be installed at the factory or added later. Call Teledyne Instruments sales for pricing and
availability.
The standard configuration of the current loop option is 4 – 20 mA. 0-20 mA is also available.
5.3. Ozone Destruct Option
measuring O3 levels that are ≥100 ppb. Ozone levels this high will damage several of the
components of the monitor downstream of the sensor module such as the flow meter and the
pump. High levels of O3 are hazardous and should be removed from the gas stream exhaust.
CAUTION
Make sure that the exhaust line is vented to an outside area.
05228 Rev B
DCN 5164
23
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
Optional Hardware and Software
460L Instruction Manual
USER NOTES:
24
05228 Rev B
PRINTED DOCUMENTS ARE UNCONTROLLED
DCN 5164
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
460L Instruction Manual
Operating Instructions
6. OPERATING INSTRUCTIONS
The 460L has been designed for simple and trouble-free operation. The sections below detail the
operational features of the 460L.
6.1. Front Panel Display
6.1.1. O readout
3
The current ozone concentration is displayed in the 4-digit readout in the center of the display.
The concentration is displayed in the currently selected units either: wt%, g/Nm3 or ppm, ppb.
6.1.2. Zero Point Calibration
The 460L needs little or no calibration in the field; however sometimes minor measurement
offsets can occur. To compensate for this a zero point calibration can be initiated by using the
two ZERO buttons. See Section 8.1 for more information.
6.1.3. Status LED’s
The four status LED’s to the right of the display indicate the general status of the 460L Monitor.
During normal operation, after the monitor has warmed up, the green ‘Sensor OK’ LED should be
on and all other Status LED’s should be off. For information on troubleshooting using the Status
LED’s, see Sections 11.2.4.
Table 6-1 Status LED’s
NAME
ON STATE
OFF STATE
OFF when
Reference or Measure > 4995mV;
SENSOR O.K.
Normal State is STEADY GLOW
Reference < 1000mV
STEADY GLOW if:
Ozone Pressure > 14.9 psia,
Negative Ozone Concentration,
Concentration Over-Range
INVALID
READING
Normal State
STEADY GLOW if Reference
Detector<2500mV
LAMP LOW
Normal State
Normal State
BLINKS whenever the measured O3
concentration exceeds either the alarm
limits.
ALARM
ACTIVE
Definitions for these four LED’s correspond to the definitions of the monitor’s four digital status
05228 Rev B
25
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
Operating Instructions
460L Instruction Manual
6.2. Concentration Alarms
There are two O3 concentration alarms, HI and HI-HI. Both alarms are triggered when the
measured O3 concentration equals or rises above the set limit for that alarm. The set limit for
HI alarm must always be a lower value than that of the HI-HI alarm. If the O3 concentration is
between the HI limit and the HI-HI limit, the HI alarm will be active, but the HI-HI alarm will be
inactive.
When either the HI or the HI-HI alarms are active the ALARM ACTIVE status LED on the front
panel (see Figure 3-1) will blink. In addition when the HI alarm is active alarm trigger LED D7 (on
the main PCA, see Figure 3-12) will glow and when the HI-HI alarm is active alarm trigger LED
D8 will glow.
6.2.1.1. Concentration Alarm Configuration
Either of the concentration alarms can be independently configured via the 460L’s serial
communication port (see Section 7.3.11).
The user can set the alarms to operate in either the latching or non-latching as well as
independently adjust the trigger levels of the alarm limits.
In non-latching mode, the alarms will be triggered when O3 concentration equals or exceeds the
associated limit is reached and will automatically return to an inactive state if the O3 concentration
falls below the limit level. In latching mode once an alarm is triggered it will stay active until
cleared by the user regardless of how the O3 concentration measurement changes.
6.2.1.2. Remotely Sensing the Status of the Alarm
Besides the LEDs located on the front panel and on the monitor’s main PCA, the status of each
alarm can be sensed externally via a single-pull, double-throw relay (see Section 3.5.4) or
determined from a remote location via the monitor’s serial communication port (See Chapter 7).
6.2.1.3. Clearing Alarms
When set for non-latching mode the alarms will only clear when the alarm condition disappears
and will do so automatically. When in latching mode, the concentration alarms may be cleared in
several ways:
FRONT PANEL: Pressing the ALARM ACKNOWLEDGE key on the monitor’s front panel
will clear all active alarms.
SERIAL COMMUNICATION PORT: The ALMACK command to clear all active alarms (see
Section 7.3.1)
Table 6-2 Concentration Alarm Default Settings
ALARM
HI
CONDITION
MODE
TRIGGER LIMIT
100 ppb
Enabled
Latching
HI-HI
300 ppb
26
05228 Rev B
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
460L Instruction Manual
Serial Communications
7. SERIAL COMMUNICATIONS
The 460L comes equipped with a powerful digital Serial Communications Port that can be used for
Data Acquisition and for changing the monitor’s configuration. This port can be configured for
port and connecting it to a computer or data acquisition system.
7.1. Serial Port Command Syntax
All characters sent and received are standard ASCII characters and all numbers are decimal
numbers converted to ASCII text.
All commands are sent using the following syntax:
<address><command>:<data1>,<data2>#<checksum (optional)><CR>
Where:
address
is the monitor address (default =1)
command is the command string being sent
: (colon)
is the data separator and is only included if data is being sent as part of the
command (See Command Details below to see if a command requires data or
not)
data1
data2
#
is the first data parameter, if required.
is the second data parameter, if required.
is the Checksum separator, sent only if optional checksum is included
The checksum is an ASCII checksum of all characters up to the # character.
The checksum is optional. Commands sent without the checksum (and
checksum separator,) are also valid.
CR
is a carriage return, ASCII 13.
Examples
Valid Commands with no data:
Checksum Included: 1ALMACK#474<CR>
No Checksum: 1ALMACK<CR>
Valid Commands with data:
Checksum Included: 1VSET:1,20#620<CR>
No Checksum: 1VSET:1,20<CR>
05228 Rev B
DCN 5164
27
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
Serial Communications
460L Instruction Manual
7.2. Serial Port Command Summary
Table 7-1 below lists the commands available and a summary of their function.
Table 7-1 Serial Port Command Summary
COMMAND
DESCRIPTION
ALMACK
Acknowledges and clears any active concentration
alarms
ALMSTAT
CZERO
Returns the current state of the concentration
alarms
Perform a manual zero calibration. The CZERO
does not activate the zero-gas solenoid valve,
therefore the calibration is calculated based on the
current ozone content of the measurement cell.
DACSTEP
O3
Analog Output Test Mode, Step Function
Returns O3 concentration currently being
measured
SETADDR
TDUMP
TLIST
Sets communication address for this 460L to a
specific value
Returns the current values of a set of
measurement parameters as a single data string
Returns list of measurement parameters and their
current values as a formatted list.
VGET
VLIST
VSET
Returns the current value of a single VAR
Lists all VARS and their current values
Sets value of internal VAR
28
05228 Rev B
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
460L Instruction Manual
Serial Communications
7.3. Serial Port Command Reference
7.3.1. ALMACK
SYNTAX
<address>ALMACK<CR>
DESCRIPTION
Acknowledges any active concentration alarm and clears them if possible For more information
see Section 6.2.
DATA PARAMETERS SENT
None
RESPONSE
<address>:<success_flag>#<checksum><CR>
EXAMPLE
Command:
1ALMACK<CR>
Response:
1:OK#261<CR>
- Command successfully received. All alarms capable of being cleared
have been.
NOTE
The alarm(s) will only clear if the condition causing the alarm no longer exists,
EXAMPLE
If the current O3 concentration rises above the trigger levels for both the HI and HI-HI alarms,
both alarms will activate. If the O3 level falls below the HI-HI alarm limit, but is still above the
is still above the HI limit, the ALMACK command will only clear the HI-HI alarm.
05228 Rev B
DCN 5164
29
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
Serial Communications
460L Instruction Manual
7.3.2. ALMSTAT
SYNTAX
<address>ALSTAT<CR>
DESCRIPTION
Returns the current status of both of the concentration alarms.
0= Inactive
1= Active
DATA PARAMETERS SENT
None
RESPONSE
<address>:<HI alarm state>,<HI-HI alarm state>#<checksum><CR>
EXAMPLE
Command:
1ALMSTAT<CR>
Response:
1:0,0#247<CR>
1:1,0#248<CR>
1:1,1#249<CR>
- Both concentration alarms are INACTIVE
- The HI alarm is ACTIVE and the HI-HI alarm is INACTIVE
- Both concentration alarms are ACTIVE
30
05228 Rev B
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
460L Instruction Manual
Serial Communications
7.3.3. CZERO
SYNTAX
<address>CZERO<CR>
DESCRIPTION
Performs a zero calibration using gas sourced from the ozone gas inlet.
DATA PARAMETERS SENT
None
RESPONSE
<address>:<success_flag>#<checksum><CR>
EXAMPLE
Command:
1CZERO<CR>
Response:
1:OK#261<CR>
- Calibration Successful
- Calibration Failed
1:FAIL#391<CR>
NOTE
Once the CZERO function is activated the monitor will briefly display dashes (‘----‘) after which
the concentration should quickly go to zero.
The CZERO does not activate the zero-gas solenoid valve, therefore the gas flowing through the
monitor is sourced from the ozone inlet and the zero calibration is calculated based on the current
ozone content of that gas source.
Care must be taken to ensure that all ozone is purged from the monitor before activating the
CZERO function. Before activating the CZERO function, disconnect the O3 supply line from the
monitor and allow room air to flow through the monitor for several minutes.
Use a shutoff valve to make sure that the O3 source does not continue to feed O3 into the supply
line while it is disconnected
05228 Rev B
DCN 5164
31
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
Serial Communications
460L Instruction Manual
7.3.4. DACSTEP
SYNTAX
<address>DACSTEP<CR>
DESCRIPTION
Puts the monitor into Analog Output setup mode. The Analog Output steps from zero to full-scale
in 25% increments, pausing for 10 seconds at each level. This repeats 5 times, after which the
monitor returns to normal operation. This mode is useful for testing the Analog Output and the
operation of any equipment measuring the Analog Output.
DATA PARAMETERS SENT
None
RESPONSE
<address>:<success_flag>#<checksum><CR>
EXAMPLE
Command:
1DACSTEP<CR>
Response:
1:OK#261<CR>
NOTES
- DACSTEP command acknowledged and initiated
The DACSTEP function takes some time to complete.
When the command is sent to the monitor, it will immediately respond with the <address> and
colon ‘:’ as an acknowledgement that the message was received. After the function is complete
the rest of the response will be sent.
No additional commands should be issued to the monitor until the function completes.
32
05228 Rev B
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
460L Instruction Manual
Serial Communications
7.3.5. O3
SYNTAX
<address>O3<CR>
DESCRIPTION
Returns the current ozone concentration measured by the monitor.
DATA PARAMETERS SENT
None
RESPONSE
<address>:<o3_conc>#<checksum><CR>
EXAMPLE
Command:
1O3<CR>
Response:
1:250.1898#522<CR> - Current O3 Concentration (reading 250.2 ppb)
NOTES
While the concentration value returned shows more digits after the decimal than the front panel
(in the example above, 250.1898) display it is only valid to 4 significant digits (in the example
above, 250.2).
The returned value will be in whichever units of measure for which the monitor is set.
05228 Rev B
33
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
Serial Communications
460L Instruction Manual
7.3.6. SETADDR
SYNTAX
<address>SETADDR:<new_address><CR>
DESCRIPTION
Changes the communications address to a new value.
DATA PARAMETERS SENT
new_address is the new address for the monitor; Range for new_address is 1-9.
RESPONSE
<address>:<success_flag>#<checksum><CR>
EXAMPLE
Command:
1SETADDR:2<CR>
Response:
- Change address from 1 to 2
1:OK#261<CR>
- Change Address Successful
- Change Address Failed
1:FAIL#391<CR>
NOTES
The monitor response is from the previous address (1 in the example shown above), but any
further commands must be at new address (2 in the example shown above).
If the address change was successful, after sending back the OK response, the monitor will no
longer respond to commands with address 1.
34
05228 Rev B
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
460L Instruction Manual
Serial Communications
7.3.7. TDUMP
SYNTAX
<address>TDUMP<CR>
DESCRIPTION
This command is best used if the response is intended as input for a database program or data
acquisition system
It returns a string made up of the current values of the following parameters in the following
order: O3 Concentration, Cell Pressure (psia), Cell Temperature (K,) Lamp Temperature (K,)
Measure Detector (mV,) Calibrated Reference Detector (mV,) Reference Detector (mV).
DATA PARAMETERS SENT
None
RESPONSE
<address>:<o3_conc>,<pressure>,<cell_temp>,<lamp_temp>,<measure>,<cal_ref>,<referenc
e>,<al-hi-stat>,<al-hi-hi-stat>#<checksum><CR>
EXAMPLE
Command:
1TDUMP<CR>
Response:
1:0.0282144,14.77461,300.7179,324.7713,2881.437,2940.903,4412.52,1,0#3413
<CR>
NOTES
While the concentration value returned shows more digits after the decimal than the front panel
display (in the example above, 0.0282144), it is only valid to 4 significant digits (in the example
above, 0.028).
05228 Rev B
DCN 5164
35
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
Serial Communications
460L Instruction Manual
7.3.8. TLIST
SYNTAX
<address>TLIST<CR>
DESCRIPTION
Returns a formatted, easy to read list of the parameters by name with the current values for
each.
DATA PARAMETERS SENT
None
RESPONSE
Test Parameter List (See Below)
EXAMPLE
Command:
1TLIST<CR>
Response:
O3 = 0.0226168
Press = 14.7753
Cell Temp = 300.7116
Lamp Temp = 324.7965
Ref = 2881.52
Meas = 2941.092
Raw Ref = 4412.646
HI Alarm = ON
HI-HI Alarm = OFF
NOTES
The set of parameters is not the same as those returned by the TDUMP command.
While the concentration value returned shows more digits after the decimal than the front panel
display (in the example above, 0.0226168), it is only valid to 4 significant digits (in the example
above, 0.022).
No checksum is sent in the response to the TLIST command.
36
05228 Rev B
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
460L Instruction Manual
Serial Communications
7.3.9. VGET
SYNTAX
<address>VGET:<var_index><CR>
DESCRIPTION
Returns value of an internal configuration variable (VAR.)
DATA PARAMETERS SENT
var_index
is index number for internal VAR as follows:
Table 7-2 VAR_INDEX List for VGET Command
var_index
Name
Description
ANALOG_RANGE
Full-Scale concentration range for Analog Output scaling
0
1
0 = disabled
1 = enabled
ALARM_ENABLE
ALARM_MODE
0 = latching
1 =non-latching
2
Molecular weight of carrier gas(e.g. 32.0 = O2). Only used by
460L’s with ppmw capability
CARRIER_WEIGHT
COMM_MODE
IIR_FILT
3
4
5
Not Used
The sensitivity of the software filter used by the monitor to reduce
noise and hysteresis in the reported O3 concentration reading.
2 =ppb
3 = ppm
CONC_UNITS
6
HI_ALARM_LEVEL
Sets the trigger limit of the HI alarm
Sets the trigger limit of the HI-HI alarm
7
8
HI-HI_ALARM_LEVEL
NOTE: Only one yar_index is allowed per iteration of the VGET command
RESPONSE
<address>:<var_value>#<checksum><CR>
EXAMPLE
Command:
1VGET:8<CR> - Request limit value for HI concentration Alarm
Response:
1:300.0#348<CR>
NOTE
- HI alarm limit is 300.0 ppb
Response is in units of measure for which the monitor is set.
05228 Rev B
37
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
Serial Communications
460L Instruction Manual
7.3.10. VLIST
SYNTAX
<address>VLIST<CR>
DESCRIPTION
Returns a formatted, easy to read list of an internal configuration variables (VAR’s) and their
current values.
DATA PARAMETERS SENT
None
RESPONSE
VAR List (See Below)
EXAMPLE
Command:
1VLIST<CR>
Response:
#0 analog_range = 1000.0
#1 alarm_enable = 1.0
#2 alarm_mode = 0.0
#3 carrier_weight =32.0
#4 comm_mode = 0.0
#5 iir_filt = 0.25
#6 conc_units = 2.0
#7 hi_al_level = 100.0
#8 hihi_al_level = 300.0
NOTES
No checksum is sent in the response to the VLIST command.
The 460L does not use the comm_mode VAR, so its value will always be 0.
hi_al_level and hihi_al_level are returned in the units of measure for which the monitor is set.
38
05228 Rev B
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
460L Instruction Manual
Serial Communications
7.3.11. VSET
SYNTAX
<address>VSET:<var_index>,<new_value><CR>
DESCRIPTION
Sets value of an internal configuration variable (VAR.)
DATA PARAMETERS SENT
var_index
new_value
index number for internal VAR (See VGET for index list)
new value for VAR.
RESPONSE
<address>:<var_value>#<checksum><CR>
Table 7-3 VAR_INDEX List for VSET Command
var_index
Name
Description
Allowable Range
Sets the full-scale concentration range for
Analog Output scaling
1 ppb – 1000 ppb
0.001 ppm – 1.000 ppm
ANALOG_RANGE
0
0 = disabled
1 = enabled
ALARM_ENABLE
ALARM_MODE
Enables or disables the Concentration alarms
1
2
Selects latching mode or non-latching mode for
both alarms.
0 = latching
1 =non-latching
Sets the molecular weight of carrier gas for
wt% calculations (e.g. 32.0 = O2)
CARRIER_WEIGHT
COMM_MODE
27.0 – 32.0
N/A
3
4
Not Used
Sets the sensitivity of the software filter used
by the monitor to reduce noise and hysteresis
in the reported O3 concentration reading.
0.05 – 1.0
IIR_FILT
5
2 =ppb
3 = ppm
CONC_UNITS
HI_AL_LEVEL
HIHI_AL_LEVEL
O3 concentration measurement units
Sets the trigger limit of the HI alarm
Sets the trigger limit of the HI-HI alarm
6
7
8
10 ppb < x < 1000 ppb
0.010 ppm < x < 1.000 ppm
10 ppb < x < 1000 ppb
0.010 ppm < x < 1.000 ppm
NOTE: Only one yar_index is allowed per iteration of the VSET command
EXAMPLE1
Command:
1VSET:8,275.0<CR> - Set HI-HI concentration alarm to 275.0 ppb
Response:
1:OK#261<CR>
- VSET Successful
1:FAIL#391<CR>
- VSET Failed
05228 Rev B
DCN 5164
39
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
Serial Communications
460L Instruction Manual
EXAMPLE2
Command:
1VSET:6,1<CR>
Response:
- Set CONC_UNITS VAR to ppm
1:OK#261<CR>
- VSET Successful
1:FAIL#391<CR>
- VSET Failed
NOTES
ANALOG_RANGE: Setting is in whatever units selected by the CONC_UNITS VAR
ALARM_ENABLE: The concentration alarms cannot be enable/disabled independently.
ALARM_MODE: The mode of the concentration alarms cannot be set independently.
CARRIER_WEIGHT: This variable is only used by 460L’s with ppm by weight capability (ppmw).
The molecular weight of pure O2 is 32.
The nominal molecular weight of ambient air is 28.96
COM_MODE: The 460L does not use the comm_mode VAR, so its value will always be 0.
IIR_FILT: The lower the setting value the more significant the effect of the filter; 1.0 = No
filtering; 0.05 = Maximum filtering. 0.0 is not allowed.
HIHI_AL_LEVEL must be greater than HI_AL_LEVEL. New_value should be in the units of
measure the monitor for which the monitor is set.
EXAMPLE:
1VSET:8,275.0<CR>
1VSET:8,0.275<CR>
- Set HI-HI concentration alarm to 275.0 ppb
- Set HI-HI concentration alarm to 0.275 ppm
USER NOTES:
40
05228 Rev B
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
460L Instruction Manual
Calibration
8. CALIBRATION
8.1. Zero Point calibration
Variations in ambient conditions, most notably changes in ambient humidity, and minor changes
in the performance of certain electronic components over time may cause slight offsets in the zero
point measurement of the monitor. To check for and compensate for these minor offsets use the
following procedure.
1. Supply zero air to the monitor by inserting a O3 scrubber filled with activated charcoal into
the ozone supply gas line just before it enters the monitor.
A scrubber of this type is available from Teledyne Instruments
(Carbon Filter, DAU, 000 Grade; P/N FL0000020).
NOTE
Since variations in humidity is a common cause of these types of offsets, use of bottled
O2, which is extremely dry, as zero air is not recommended particularly in areas where
the ambient humidity is high.
2. Once the scrubber is installed, allow the monitor to operate until the O3 reading stabilizes.
This may take approximately 15 minutes
3. Check the O3 reading.
If the reading is within the repeatability specification of the monitor (see Section 2.1),
no adjustment is needed.
If the reading is higher than the repeatability specification of the monitor but still
relatively minor (10 to 20 ppb for example), press the two ZERO keys on the front
panel and hold them for 3 seconds.
If the offset is greater than 30 ppb or pressing the zero key has no effect, contact
Teledyne Instruments’ customer service department.
8.2. Span Point Calibration
1. Connect the M40M to a computer running a terminal emulation program such as Hyper-
2. Flow zero air, or pure O2 through the analyzer and wait 15minutes.
3. After the analyzer has stabilized, perform a zero calibration by pressing both zero buttons
on the front panel at the same time.
4. Type:
<address>LOGIN:929<CR>
05228 Rev B
DCN 5164
41
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
Calibration
460L Instruction Manual
Where
<address> is the network address of the M460M being calibrated.
929 is a special password allowing you to calibrate the analyzer.
5. List the current values of the M460M’s VARS. Type:
<address>VLIST<CR>
Look through the list for one with the a special VARS named either O3 SLOPE, or O3
SPAN SLOPE. This VAR is only available using the 929 password.
Record both the slope value and the VAR Number it corresponds with.
6. Make sure your ozone standard and analyzer are both reading the same span gas or ozone
concentration.
7. Calculate the slope using the following equations:
Equation 8-1
(O3 concentration on analyzer) ÷ (O3 Slope) = (True O3 value of analyzer)
Equation 8-2
(O3 concentration on standard) ÷ (True O3 value of analyzer) = (New slope)
EXAMPLE
Where:
Slope of analyzer = 1.013
Current O3 reading of analyzer = 8.70%
Current O3 reading of standard = 9.00%
a) (O3 concentration on analyzer) ÷ (O3 Slope) = (True O3 value of analyzer
8.70% 1.013 8.59%
b) (O3 concentration on standard) ÷ (True O3 value of analyzer) = (New slope)
9.00% 8.59% 1.048
1.048 would be the New Slope Value that needs to be entered in the next step.
÷
=
÷
=
42
05228 Rev B
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
460L Instruction Manual
Calibration
8. Enter the New Slope into the M460M’s memory enter it in. Type:
<address>VSET:<VAR Number>,<New Slope Value><CR>
EXAMPLE
3VSET:16,1.048<CR>
Where:
The address of the M460M is 3
The O3 SLOPE VAR is number 16
The New Slope is = 1.013
8.3. Adjusting the Optional Current Loop Output
If your monitor includes the option current loop output you may need to check or adjust the
actual current levels of the output to ensure that it matches the input requirements of your
recording device. See Section 3.5.1 for details on making connections to the 4-20mA output.
To manually adjust the zero and span points of the 4-20mA analog output:
1. Disconnect the monitor from AC power.
2. Connect current measuring meter in series with the 4-20mA output. For best results, the
4-20mA output should be calibrated with the actual load (measuring device) attached. If
this cannot be done, then a 250 – 500 ohm resistor should be placed in series with the
current meter to simulate a load.
See Figure 3-6 for
pin assignments of
mA
the signal I/O
connector on the
rear panel.
IN
OUT
ANALOG OUT
I IN +
I IN -
ANALOG OUT -
Recording
Device
Monitor
Figure 8-1 Setup for Measuring Current Output Signal Level
3. While reconnecting the monitor to AC power, press and hold the “Alarm Reset” button on
the front panel. This will cause the monitor to enter the analog output step mode. The
display on the monitor will display “A 0” indicating that it is in the analog output step
4. At this point the analog output should read somewhere near 4.0mA. Adjust the “Zero”
potentiometer on the 4-20mA PCA (See Figure 8-2) as necessary.
05228 Rev B
DCN 5164
43
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
Calibration
460L Instruction Manual
5. The monitor will automatically generate the nominal signal level by producing a 25%
increment output level on the current loop.
EXAMPLE
OUTPUT LEVEL
NOMINAL SIGNAL
LEVEL
0%
25%
50%
75%
100%
4mA
8mA
12mA
16mA
20mA
6. The display will then show “A100” (four times). Adjust the “Span” potentiometer (See
Figure 8.2) as necessary.
Figure 8-2 Location Current Output Adjustment Potentiometers
7. Note that the zero and span adjustments are not completely independent and adjusting
one point may slightly affect on the other. Therefore steps 4-6 may need to be repeated
several times in order to properly adjusted both points.
8. When the adjustment process is complete, the monitor will automatically restart in
standard measurement after 5 cycles.
USER NOTES:
44
05228 Rev B
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
460L Instruction Manual
Maintenance
9. MAINTENANCE:
9.1. Replacing the Particulate Filter Element
The 460L is equipped with a particulate filter on the ozone inlet. Only Teledyne Instruments filter
elements should be used. When the monitor is first installed, the sample filters should be checked
at least once a week for particulate loading and replaced if necessary. Once the replacement
frequency is determined, a regular schedule for filter replacement should be instituted.
For replacement filter, please contact Teledyne Instruments’ sales department and request part
number 05017.
9.1.1. Filter Replacement Procedure
1. Turn the Monitor off and ensure that the gas delivery line is not under pressure
2. Purge the gas delivery line of ozone.
3. Unscrew the pressure fitting at the top of the filter (Fitting B in Figure 9-1) from the union
fitting.
4. Unscrew the pressure fitting at the bottom of the filter (Fitting C in Figure 9-1) from the
ozone Inlet.
5. Discard the filter.
6. Attach the new filter using the fittings supplied with it.
7. After reassembly, the gas line should be pressurized with oxygen or dry air and checked
for leaks using a bubble solution.
Pressure Fitting A
Pressure
Fitting B
Union
Fitting
Pressure
Fitting C
Filter
Ozone Inlet
Figure 9-1 Changing the Particulate Filter
05228 Rev B
DCN 5164
45
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
Maintenance
460L Instruction Manual
9.1.2. Mounting the Particulate Filter Externally
In applications where the 460L is sampling ambient air at its own location and no inlet gas line is
being used, the particulate filter can be mounted on the outside of the instrument so that it may
be replaced without turning off the monitor or opening the case.
To mount the particulate filter outside the monitor enclosure:
1. Turn the monitor off and ensure that the gas delivery line is not under pressure
2. Purge the gas delivery line of ozone.
3. Unscrew the pressure attached to the internal gas line (Fitting A in Figure 9-2) from the
union fitting.
4. Unscrew the union fitting from the fitting at the top of the filter (fitting B in Figure 9-2).
Although the union fitting is no longer needed, save it in case you wish to return the
particulate filter to the inside of the monitor at a later date.
5. Unscrew the pressure fitting at the bottom of the filter (Fitting C in Figure 9-2) from the
ozone inlet.
6. Reattach the filter to the external side of the ozone inlet.
7. Attach the internal gas line to the internal side of the ozone inlet using Fitting A.
8. After reassembly, the gas line should be pressurized with oxygen or dry air and checked
for leaks using a bubble solution.
AFTER
BEFORE
Pressure
Fitting A
Union
Fitting
Pressure
Fitting B
Pressure
Fitting C
Pressure
Fitting A
Filter
Pressure
Fitting B
Ozone Inlet
Ozone Inlet
Inlet Gas Line
Filter
Pressure
Fitting C
No Inlet Gas Line
Figure 9-2 Changing the Particulate Filter
46
05228 Rev B
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
460L Instruction Manual
Maintenance
9.2. Maintaining the Optional H2O Coalescing Filter
9.2.1. Draining the Optional Coalescing Filter
The coalescing filter component of the optional sample conditioning system may accumulate with
water (see Figure 9-3 for filter location). It must be checked periodically and drained.
It is recommended that the filter be checked every 2 hours for the first several days of the
monitor’s operation to determine the fill rate of the filter’s reservoir.
To drain the filter:
1. Disconnect ozone supply line from monitor and shut off flow to the monitor using needle
valve on front panel.
2. Remove the cap from the fitting on the bottom of the coalescing filter and allow the filter to
drain.
3. Replace cap.
4. Reconnect the O3 supply line and adjust flow to the monitor.
5. Open the flow meter valve and adjust the gas flow to the appropriate rate.
Coalescence
Filter
(optional)
Coalescence
Filter Drain
Figure 9-3 Draining the Optional H2O Coalescing filter
05228 Rev B
DCN 5164
47
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
Maintenance
460L Instruction Manual
9.2.2. Replacing the Coalescing Membrane
The coalescing filter contains a replaceable Teflon® coalescing membrane. Under normal
operating conditions, this membrane should last for a long time, however a significant reduction in
the gas pressure measured in the O3 cell might indicate that this membrane needs replacement.
To replace the filter membrane:
1. Disconnect ozone supply line from the monitor and shut off flow to the monitor using
needle valve on front panel.
2. Remove inlet and exit fittings from scrubber body and cap.
3. Remove the orange cap from the top of the coalescing filter.
4. Remove the old membrane. It is located on the underside of the cap assembly
(see Figure 9-4).
5. Insert the new membrane into the cap assembly.
6. Tightly secure the cap back on the filter.
7. Reconnect the O3 supply line adjust flow to the monitor.
8. Open the flowmeter valve and adjust the gas flow to the appropriate rate.
Figure 9-4 Replacing the Membrane of the Optional H2O Coalescing filter
Replacement filter elements can be ordered from TAPI (P/N 036750000).
48
05228 Rev B
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
460L Instruction Manual
Maintenance
9.3. UV Lamp Replacement
1. Disconnect power from the monitor and open cover.
2. Loosen, but do not remove, the two lamp set-screws (see Figure 9-5 below).
3. Unplug lamp from the UV lamp power connector on Sensor Module PCA.
4. Remove lamp from lamp housing. Dispose of lamp in accordance with local regulations
regarding disposal of Mercury containing waste.
5. Install new lamp in housing and plug into the UV lamp power connector.
6. Connect the ground lead of a voltmeter to TP4.
7. Connect the positive lead of voltmeter to TP10.
8. Reconnect power to the monitor.
9. The voltage on TP10 must read between –0.6 and –1.6Vdc. If this voltage is outside that
range, slowly rotate the lamp until the proper voltage is achieved, then tighten the lamp
set-screws.
10.Connect the positive lead of a voltmeter to TP11 (Measure Detector Voltage).
11.Adjust R26 until the voltage on TP11 is as high as possible within the range of 0.60 –1.00
volts.
UV Lamp Set Screws
UV Lamp
UV Lamp
Power
Connector
Reset
button
Test Point
4
Test Point
11
Test Point
10
R26
Figure 9-5 UV Lamp Set Screws and Calibration Test Points
05228 Rev B
DCN 5164
49
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
Maintenance
460L Instruction Manual
9.4. Cleaning Exterior Surfaces of the 460L
If necessary, the exterior surfaces of the 460L can be cleaned with a damp cloth. Do not attempt
to clean any of the other surfaces of the monitor. Do not submerge any part of the monitor in
water or cleaning solution.
9.5. Degree of Protection
The Model 460L has a water ingress rating of IP65 which indicates that it can withstand strong
jets of water and is totally protected against dust.
USER NOTES:
50
05228 Rev B
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
460L Instruction Manual
Theory of Operation
10. THEORY OF OPERATION
10.1. Basic O3 Measurement Principle
10.1.1. (Beer’s Law)
The detection of ozone molecules in a gas is based on the principle that ozone is a very strong
absorber of UV light with a wavelength of 254 nm. If the distance that this light travels through
the gas is always the same, the more ozone present in a gas, the more UV light is absorbed. If
the distance the light travels through the gas, the intensity of light passing through the ozone
containing gas, as well as the intensity of the light which does not pass through the gas are all
known, the amount of ozone present can be calculated according to the following equation, called
Beer’s Law (also referred to as the Beer-Lambert equation).
I=IO e-
LC
Equation 10-1
Where:
Io
I
is the intensity of the light if there was no absorption.
is the intensity with absorption.
L
is the absorption path, or the distance the light travels as it is being absorbed. This
distance determines how many molecules are present in the column of gas in the
absorption cell.
C
is the concentration of the absorbing gas. In the case of the Model 460L, Ozone (O3).
α
is the absorption coefficient absorption coefficient, a number that reflects the inherent
ability of ozone to absorb 254 nm light. Most current measurements place this value at
-1
-1
308 cm atm at Standard Temperature and Pressure (STP). The value of this
number reflects the fact that ozone is a very efficient absorber of UV radiation which is
why stratospheric ozone protects the life forms lower in the atmosphere from the
harmful effects from solar UV radiation.
05228 Rev B
DCN 5164
51
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
Theory of Operation
460L Instruction Manual
To solve this equation for C, the concentration of the absorbing gas (in this case O3), the
application of a little algebra is required to rearrange the equation as follows:
I
o
1
C ln
I
L
Equation 10-2
Unfortunately, both ambient temperature and pressure influence the density of the sample gas
and therefore the number of ozone molecules present in the absorption path thus changing the
amount of light absorbed.
In order to account for this effect the following addition is made to the equation:
I
1
L
Τ
14.695psi
o
C ln
273o
I
Ρ
Equation 10-3
Where:
T
= sample ambient temperature in degrees Kelvin
= ambient pressure in pound per square inch (psi),
P
Finally, to convert the result into Parts per Million (PPM), the following change is made:
6
I
10
14.695psi
o
C ln
273o
I
L
Equation 10-4
10.1.2. The Absorption Path
In the most basic terms, the Model 460L uses a high energy, mercury vapor lamp to generate a
beam of UV light. This beam passes through a window of material specifically chosen to be both
non-reactive to O3 and transparent to UV radiation at 254nm and into an absorption tube filled
with Sample Gas.
Because ozone is a very efficient absorber of UV radiation the Absorption Path Length required to
create a measurable decrease in UV intensity is short enough (approximately 16 cm) that the light
52
05228 Rev B
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
460L Instruction Manual
Theory of Operation
beam is only required to make pass through the Absorption Tube. Therefore no complex mirror
system is needed to lengthen the effective path by bouncing the beam back and forth.
Finally, the UV then passes through a similar window at the other end of the Absorption Tube and
is detected by a specially designed vacuum diode that only detects radiation at or very near a
wavelength of 254nm. The specificity of the detector is high enough that no extra optical filtering
of the UV light is needed.
The detector reacts to the UV light and outputs a voltage that varies in direct relationship with
the light’s intensity. This voltage is digitized and sent to the instrument’s CPU to be used in
computing the concentration of O3 in the absorption tube.
Window
Window
UV Detector
ABSORPTION TUBE
Sample Gas IN
Sample Gas OUT
UV
Source
Absorption Path Length = 16 cm
Figure 10-1 O3 Absorption Path
10.1.3. The Reference / Measurement Cycle
In order to solve the Beer-Lambert equation it is necessary to know the intensity of the light
passing through the absorption path both when O3 is present and when it is not. The Model 460L
accomplishes this by alternately passing the sample gas through a chemical scrubber that
removes any O3 present and sending it directly to the absorption tube.
Measure Path
Reference/
Measure
Valve
(I)
Particulate
Filter
From Sample
Port
Reference Path
O3
Scrubber
(I0)
Valve switches
every 3 seconds
PUMP
To Exhaust
Port
ABSORPTION TUBE
Figure 10-2 Reference / Measurement Gas Cycle
05228 Rev B
DCN 5164
53
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
Theory of Operation
460L Instruction Manual
The Measurement / Reference Cycle consists of:
Table 10-1 Measurement / Reference Cycle
CYCLE
STATUS
VALVE STATE
(FIG. 10-3)
DURATION
- -
ACTIVITY
Measure/Reference Valve Opens to the Measure Path.
Wait Period. Ensures that the Absorption tube has been adequately
flushed of any previously present gasses.
MEASURE
PERIOD
0.50 sec.
1 3
Analyzer measures the average UV light intensity of O3 bearing
Sample Gas (I) during this period.
0.15 sec.
- -
Measure/Reference Valve Opens to the Reference Path.
Wait Period. Ensures that the Absorption tube has been adequately
flushed of O3 bearing gas.
REFERENCE
PERIOD
0.50 sec.
2 3
Analyzer measures the average UV light intensity of Non-O3 bearing
Sample Gas (I0) during this period.
0.15 sec.
1 min 30 sec
TOTAL CYCLE TIME
10.1.4. Digital Noise Filter
The 460L software processes sample gas concentration data through a noise filter that stabilizes
the concentration value reported to the display and via the monitor’s analog outputs.
54
05228 Rev B
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
460L Instruction Manual
Theory of Operation
10.2. Pneumatic Theory of Operation
10.2.1. Basic Pneumatic Flow And Flow Control
Monitor Enclosure
Flow Meter Valve
PUMP
Exhaust
Out
O3 Measurement Cell
Reference Cycle Gas Path
MEASURE
REFERENCE
VALVE
O3
Scrubber
3
2
O3 IN
1
Measure Cycle Gas Path
Gas
Pressure Sensor
O3
DESTRUCT
(Optional)
Figure 10-3 460L Internal Pneumatic Diagram – Basic Configuration
10.2.2. Internal Pump and Flow Control
Air flow through the M460L O3 Monitor is supplied by a single-diaphragm, brushless DC pump that
pulls air though the monitor. Since diaphragm pumps necessarily heat and compress the air they
are pumping and since both temperature and pressure fluctuations can effect the O3
measurement, the pump is placed down stream from the measurement cell to avoid any
inadvertent effects resulting from the pumping action.
An adjustable needle-restrictor valve and flow gauge, located on the front panel of the monitor
allow the user to manually adjust the gas flow rate through the monitor.
Particulate Filter
To remove particles in the sample gas which might clog airways or foul the measurement cell
optics, the monitor is equipped with a glass-fiber membrane filter of 25 mm diameter with a pore
size of 1.5 microns. The filter is located inside the black housing on the bottom of the monitor.
See Section 9.1 for location and instruction for replacing the filter element.
05228 Rev B
DCN 5164
55
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
Theory of Operation
460L Instruction Manual
10.2.3. Optional Sample Conditioning
The source air measured by the 460L needs to be as dry as possible. Significant amounts of liquid
or vaporous water present in the source gas can foul the measurement cell optics. Also, water
absorbs O3 and interferes with the 460L’s ability to accurately measure the O3 in the source gas.
To counteract this problem, several optional components can be added to the 460L.
Monitor Enclosure
Flow Meter Valve
PUMP
Exhaust
Out
O3 Measurement Cell
MEASURE
REFERENCE
VALVE
O3
Scrubber
O3 IN
3
2
1
Vapor
Dryer
Purge
Line
Gas
Pressure Sensor
O3
DESTRUCT
(Optional)
H2O
Coalescing
Filter
(Optional)
Figure 10-4 460L Internal Pneumatic Diagram with Optional Sample Conditioning
10.2.3.1. H2O Coalescing Filter
The first step of this drying process is to remove any liquid water from the source gas. The 460L
uses a Teflon® membrane, coalescing filter to accomplish this. This filter works in two ways. First,
droplets of water that are large enough to precipitate out of the air on their own simply fall to the
bottom of the filters container. Second, smaller droplets, small enough to stay combined and
bourn along with the air encounter the Teflon® membrane (47 mm diameter; 20 micron pore size)
at the top of the filter, and because Teflon® is inherently water repellent, these tiny droplets
collect along the Teflon® fibers combining and growing until they are large enough to drip down
into the reservoir.
10.2.3.2. H2O Vapor Dryer
Once all of the liquid water is removed from the source gas, a separate, Perma Pure® single tube
permeation tube dryer removes any vaporous water still present. The dryer consists of a single
tube of Nafion®, a co-polymer similar to Teflon® that absorbs water very well but not other
chemicals. The Nafion® tube is mounted within an outer, flexible plastic tube. As gas flows
through the inner Nafion® tube, water vapor is absorbed into the membrane walls. The absorbed
water is transported through the membrane wall and evaporates into the dry purge gas flowing
through the outer tube, countercurrent to the gas in the inner tube. This process is called
pervaporation and is driven by the humidity gradient between the inner and outer tubes as well as
the flow rates and pressure difference between inner and outer tubing.
56
05228 Rev B
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
460L Instruction Manual
Theory of Operation
To provide a dry purge gas for the outer side of the Nafion® tube, the 460L returns the dried O3
free air from the measurement cell and O3 destruct to the outer tube (see Figure 10-4). When the
monitor is first started, the humidity gradient between the inner and outer tubes is not very large
and the dryer’s efficiency is low at first but improves as this cycle reduces the moisture in the
sample gas and settles at a minimum humidity.
10.2.3.3. Ozone Destruct Scrubber
The ozone destruct scrubber removes O3 from the exhaust gas stream of the monitor after it exits
the instrument’s absorption tube but before it can damage the 460L’s flow meter and pump. It
also prevents hazardously high levels of O3 from exiting the monitors exhaust outlet.
When installed, the O3 scrubber is located outside the monitor’s NEMA housing, on the right side
at the top (see Figure 3-5). It is filled with a special catalytic ozone scrubbing material that
removes all of the O3 from the sample gas exiting the sensor module’s absorption tube. The
catalyst used in the scrubber only converts ozone to oxygen and does not produce any toxic or
hazardous gases.
The catalyst is 100% efficient at scrubbing ozone at room temperature. It is a true catalytic
converter, therefore there are no maintenance requirements such as changing the scrubbing
material as is required for charcoal-based scrubbers.
05228 Rev B
DCN 5164
57
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
Theory of Operation
460L Instruction Manual
10.3. Electronic Theory of Operation
Front Panel
Keyboard
Status LED’s
Sensor OK
Display
Driver
Board
Invalid Reading
Lamp Low
LED Display
Alarm Active
I2C Bus
Sensor Module
RS-485
CPU Board
UV Lamp
UV Lamp
Power
Supply
CPU
Lamp
Temperature
Micro-processor
Analog
Ouput
Sensor
I2C Bus
Measurement
RS-232 / RS-485
Port
Cell
Temperature
Sensor
Low
Resolution
A/D
High
Resolution
A/D
Optional 4-20 mA
outpt
Control
Outputs
Detector Assy
Pre
Amp
Measurement
Detector
Amplifier
Sensor OK
Relay
Status
Output
Drivers
Sample Gas
Pressure
Sensor
HI Alarm Relay
M/R Valve
Control
Circuitry
HI-HI Alarm
Relay
M/R Valve
Relay
M/R Valve
Control
Inputs
Status
Output
Analog
Outputs
Serial I/O
UV Lamp
Heater
Analog and Digital I/O Connector
Main Board
Figure 10-5 460L Electronic Block Diagram
Electronically, the 460L is of modular design (see Figure 10-5). Each Sub-module performs a
specific set of functions as described in Sections 10.3.1 through 10.3.5.
10.3.1. Main Board
This printed circuit assembly provides interconnection between the monitor’s other electronic
modules; some opto-isolated signal buffers for the digital status outputs and control inputs and is
the location of the three solid state output relays (see Section 3.5.5).
The monitor’s main power supply (see Section 10.3.5) is also located on this assembly.
58
05228 Rev B
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
460L Instruction Manual
Theory of Operation
10.3.2. O3 Sensor Module
The heart of the 460 Monitor is the O3 sensor module. This electromechanical assembly located at
the left hand side of the enclosure includes all of the pneumatic, mechanical and electronic
components needed to gather the data required to calculate the O3 content of the source gas.
10.3.2.1. O3 Sensor Components
UV Lamp: The ultraviolet light needed to detect O3 is supplied by a mercury-vapor UV
lamp. This lamp is coated in a material that optically screens the UV radiation output to
remove the O3 producing 185nm radiation. Only light at 254nm is emitted.
UV Lamp Heater : to operate efficiently the UV lamp must be kept at a temperature of
52˚C or higher. While the heat created by the lamp itself is usually sufficient to cause this,
under some ambient conditions additional heating is required. This additional heat is
provided by a DC heater, controlled by the sensor microprocessor.
Temperature Sensors: Two solid state temperature sensors are located in the O3 sensor
module. They are:
Measurement Cell Temperature Sensor: This sensor detects the temperature of the gas
inside the measurement cell. This information is used by the CPU as part of the O3
concentration calculation (see Formula 10-3 in Section 10.1).
UV Lamp Temperature Sensor: This sensor, attached to the UV lamp reports the
current temperature of the Lamp to the sensor microprocessor via the sensor module
A/D converter.
Both Sensors have built-in A/D converters and send digitized data directly to the monitors
CPU.
UV detector: A UV detector measures the two primary variables, I and I0 (See Section
10.1.1) needed to compute the O3 concentration of the source gas. They are:
The first measurement (I) is taken during the measure period of the measure/
reference cycle (see Table 10-1) and records the intensity of the UV light passing
through the O3 bearing source gas.
The second measurement (I0) occurs during the reference period of the cycle and
records the intensity of the light passing though gas from which the O3 has been
removed.
This detector is a specially designed vacuum diode that only reacts to radiation at or very
near a wavelength of 254nm and outputs a voltage that varies in direct relationship with
the light’s intensity. The wavelength specificity of the detector is high enough that no extra
optical filtering of the UV light is needed.
Two stages of the preamplifier are used to amplify the output signals of the detector to a
level readable by the A/D Converter circuitry of the monitor’s sensor module. The first
stage of amplification is located on the PCA’s on which the detector itself is mounted. The
second stage of amplification is located on the sensor module PCA.
05228 Rev B
DCN 5164
59
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
Theory of Operation
460L Instruction Manual
Gas Pressure Sensor: This absolute pressure sensor measures the gas pressure in the
measurement cell upstream of the Pump. The sample pressure is used by the CPU to
calculate O3 Concentration (see Formula 10-3 in Section 10.1). This sensor outputs an
analog signal to the sensor microprocessor.
Measure / Reference Valve: This valve alternates the O3 gas stream between a direct
path to the absorption tube and a path that first passes it the O3 scrubber. The state of this
valve is the Sensor Module PCA (see Section 10.3.2.2).
10.3.2.2. Sensor Module PCA
The sensor module PCA performs the real work of operating the O3 sensor module. It gathers the
various measurements used to calculate the O3 concentration and performs all basic computations
related to determining the O3 concentration of the source gas.
It includes:
Sensor Signal A/D Conversion: The output of the monitor’s O3 sensor and the pressure
sensor are converted into digital signals that the CPU can understand by two analog to
digital converters (A/D) located on the sensor PCA.
LOW RESOLUTION A/D: A 12-bit, SAR converter that digitizes the output of the
monitors pressure sensor.
HIGH RESOLUTION A/D: This component digitizes the signal output by the UV detector
of the monitor’s measurement detector. Since the reference detector output requires a
large dynamic range and superior noise rejection, a Delta-Sigma type A/D Converter is
used here.
Sensor Microprocessor: This IC provides two important functions.
A/D Converter Signal Selection: The A/D converter can only convert one signal at a time.
The sensor microprocessor selects which of the sensor inputs is to be converted(pressure
sensor or measurement detector), starts the conversion process, stops it and extracts the
digital data.
Data Output: The sensor microprocessor collects data from the various digital sensors as
well as the data converted by its internal A/D and sends it to the main CPU via an internal
RS-485 serial data bus.
UV Lamp Heater Control: The sensor microprocessor also provides direct control of the UV
lamp heater by turning it on and off via a transistor using high frequency pulse width
modulation. The output of the UV Lamp sensor is converted by the micrprocessor’s A/D
converter. Based on this digital value the sensor microprocessor sends out pulses that
turn the heater ON/OFF. The more heat needed the longer the width of the ON (logic high)
pulse and the shorter the width of the OFF (logic low) portion of the pulse.
Measure / Reference Valve Control: The Measure/Reference valve is actuated by a 12
VDC solenoid valve driver that is controlled by timing circuitry built into the sensor module
board.
60
05228 Rev B
PRINTED DOCUMENTS ARE UNCONTROLLED
DCN 5164
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
460L Instruction Manual
Theory of Operation
10.3.3. CPU Board
In addition to being responsible for all I/O functions and managing the state of the monitors
various status indicators and alarms (see below), the 460L’s CPU also calculates the monitor’s
offset during zero point calibration (see Section 8.1) and converts O3 concentration data from ppb
units into ppm units if necessary.
10.3.3.1. I/O Functions
Display Data: Two-way communications between the CPU and the Display driver module
is handled via an I2C interface. I2C is a two-wire, clocked, bi-directional, digital serial I/O
bus that is used widely in commercial and consumer electronic systems.
Keyboard Input: The three keys/buttons on the front panel (two zero keys and the
Alarm Resert key) are sensed directly by the CPU as simple digital contact closures.
Analog Output: The 460L is equipped with one analog output which reports the current
O3 concentration currently being measured by the monitor. During auto zero operation the
last valid concentration value is held until the auto zero procedure is completed. This
output is factory configurable as either a 0-5 VDC signal or a 4-20 mA signal.
Serial I/O: A standard RS-232 or RS-485 serial communications port. Section 3.5.4
describes how to configure and make connections to this port. Chapter 7 describes the
syntax and commands available for use.
Control Inputs: These inputs are used to initiate certain operations. (see Table 3-3).
They are triggered by providing a contact closure or low impedance current path between
the input and the ground pin (GND – see Figure 3-9). This can be done by using a
mechanical switch or isolated transistor output from another device, such as a PLC.
10.3.3.2. Status and Alarm Functions
Relay Outputs: The 460L is equipped with three relay three SPDT relay. They are located
at the top right hand side of the main board and are labeled RELAY 1, RELAY 2 & RELAY 3.
RELAY 1 corresponds to the Sensor OK status output and LED;
RELAY 2 corresponds to the HI concentration alarm, and;
RELAY 3 corresponds to the HI-HI concentration alarm
Status LED’s: Based on data received from the sensor module, it activates and
deactivates the four status LED’s located on the monitor’s front panel (see Section 11.2.4).
Status Outputs: Logic-Level voltages are output via optically isolated NPN transistors,
which sink up to 50 mA of DC current. They are accessed through connector J2 on the
main board. Several of these outputs convey good/bad information about key monitor
operational conditions (see Section 11.2). Others reflect the status of the monitors O3
concentration alarms.
These outputs can be used to interface with devices that accept logic-level digital inputs,
such as programmable logic controllers (PLC’s). Each Status bit is an open collector output
that can withstand up to 40 VDC. All of the emitters of these transistors are tied together
05228 Rev B
DCN 5164
61
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
Theory of Operation
460L Instruction Manual
and available at the STATUS COM pin of J2 on the main board (see Figure 3-7). This pin is
normally connected to the input ground of the external device.
NOTE
Most PLC’s have internal provisions for limiting the current that the input will
draw from an external device. When connecting to a unit that does not have this
feature, an external dropping resistor must be used to limit the current through
the transistor output to less than 50 mA. At 50 mA, the transistor will drop
approximately 1.2V from its collector to emitter.
10.3.4. Display Driver and Keyboard Assembly
10.3.4.1. Keyboard
The keyboard of the M460L is comprised of 3 contact closure button/keys that are directly sensed
by the monitor’s CPU. These Switches are:
Zero Switches: When pressed simultaneously, these switches activate the monitor’s auto
zero calibration feature (see Section 8.1). Activating either switch independently has no
affect on the monitor’s’ operation.
Pressure Switch: Pressing and holding this switch causes the monitor to display the
current gas pressure of the source gas as measured by the gas pressure sensor located on
the measurement cell. Pressure is displayed in units of psia (pounds per square inch
absolute).
10.3.4.2. Display
The main display of the monitor is a 4-digit, 7-segment LED display with decimal point. Under
normal operation it displays the current O3 concentration of the source gas. It can also
momentarily display the gas pressure of the source gas.
10.3.4.3. Display Driver
The circuitry on the display has driver several functions.
Signal levels from the three front panel key/buttons are passed through the driver
unaltered, directly to the CPU.
Under command of the CPU a control chip located on this assembly turns the four status
LED’s(see Section 112 & 11.3) ON/OFF
A bipolar integrated circuit decodes the serial data sent by the CPU via an I2C bus and the
individual segments of the display ON/OFF. The clock signal used to decode this data is
supplied by the monitor’s main CPU.
The four digits on the display are controlled by multiplexing between two pairs of 2 digits each.
The display is operated in static mode. Each value sent by the CPU is held on the LED display
until a new value is sent.
62
05228 Rev B
PRINTED DOCUMENTS ARE UNCONTROLLED
DCN 5164
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
460L Instruction Manual
Theory of Operation
10.3.5. Power Distribution
The 460L operates on 90 to 260 VAC power at either 50Hz or 60Hz. As illustrated in Figure 10-6
below, power enters the monitor via a standard 3-conductor power cord through a hole provided
in the bottom of the casing. In order to maintain the IP (NEMA4X) rating of the enclosure, an
appropriate sealed conduit connector should be used.
Serial I/O
Port
Analog
Outputs
Status I/O
Measurement
Detector
Sensor Assembly
Power Supply
Status LEDS
Sensor OK
Invalid Reading
Lamp Low
Cell Dirty
CPU
Board
± 9 VDC
Pressure
Sensor
UV LAMP
TEMPERATURE
SENSOR
+5 VDC
Main Power Supply
Main Power Supply
+ 5 VDC
MEASUREMENT CELL
TEMPERATURE
SENSOR
Keyboard
+ 12 VDC
UV Lamp
+ 15 VDC
+ 12 VDC
Power Supply
200VAC 3kHz
(1000VAC Peak)
VAC IN
UV Lamp
Heater
UV LAMP
Vacuum
Pump
Output
Relays
Drivers
M / R Valve
Display
Driver
Board
LED Display
Output
Relays
Figure 10-6 460L Power Distribution Block Diagram
MAIN POWER SUPPLY: AC line power is converted and stepped down to several DC
voltages by the main power supply:
+12 VDC: Powers the vacuum pump, the display driver and the alarm relay outputs.
+5 VDC: The basic voltage on which the CPU and logic level circuitry operates.
+15 VDC: Source voltage for the keyboard (where it is regulated down to +5 VDC),
the UV lamp heater; the measure / reference valve a secondary power supply located
on the sensor module assembly.
SENSOR MODULE POWER SUPPLY: Using +15 VDC from the main power supply, this
circuitry generates the +5, +12 & ±9 VDC supplies needed to operate its own on-board
logic devices and the various components of the O3 sensor module.
05228 Rev B
DCN 5164
63
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
Theory of Operation
460L Instruction Manual
UV LAMP POWER SUPPLY: Using +12 VDC supplied by the main power supply, this
assembly generates the 30 kHz AC voltage for the monitor’s mercury vapor UV lamp. The
output of this power supply is variable. At startup voltage level of this output can reach as
high a 1000 VAC. Once the lamp is warmed up and operating at peak efficiency, the
output should be around 200 VAC.
The 460L has no onboard ON/OFF switch. A hardwired 2 Amp fuse is located on the main power
supply assembly to provide over voltage/current protection.
USER NOTES:
64
05228 Rev B
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
460L Instruction Manual
Troubleshooting
11. TROUBLESHOOTING
This chapter gives guidelines for diagnosing system and sensor malfunctions using the five digital
Status Outputs provided by the 460L. All troubleshooting should be done after the 460L has been
turned on and allowed to warm up for at least 15 minutes.
11.1. Status Output Summary
Table 11-1 below gives a summary of the operation of the Status Outputs on the 460L. See
Section 3.5.2 Digital Status Outputs for information on connecting these outputs.
Table 11-1 Digital Status Outputs Definitions
LED on
FRONT
PANEL
OUTPUT
NAME
ON STATE
OFF STATE
#
Measure > 1230mV;
Reference > 1230mV;
UV Lamp Off (Reference < 250 mV);
Sensor
O.K.
1
YES
YES
Normal State
No data from The O3 sensor
Pressure > 14.9 psia;
Pressure < 9 psia;
Negative Ozone Concentration;
Concentration Over-Range
Invalid
Reading
2
Normal State
Reference <375 mV
3
4
Lamp Low
YES
YES
Normal State
Alarm
Active
Either the HI or the HI-HI O3
concentration alarm is active
No O3 Concentrati0on alarms are active
Measured O3 concentration is ≥
than the HI alarm set point
5
6
HI Alarm
NO
NO
Normal State
Normal State
HI-HI
Alarm
Measured O3 concentration ≥ the
HI-HI alarm set point
11.2. Troubleshooting Using Status Outputs
11.2.1. Sensor OK
The normal state for the Sensor OK output in ON. During the warm-up period on start-up this
output will stay off until the UV lamp reaches a minimum intensity. If this output remains OFF
after the 15 minute warm-up period, or goes off during normal operation the 460L is in need of
servicing.
If the Sensor OK output turns OFF
One of the analog voltages output by the sensor module (reference / measure) is too high.
This could mean that the lamp output has drifted high
The reference voltage output by the sensor module (reference / measure) is too low. This
could mean that the lamp intensity has drifted very low or the lamp is completely off.
05228 Rev B
DCN 5164
65
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
Troubleshooting
460L Instruction Manual
The O3 sensor is not responding to CPU requests for data. This could represent a variety
of electronic problems with the sensor module or internal communication between the CPU
and the sensor module.
11.2.2. Invalid Reading
The normal state for the Invalid Reading output is OFF. When this output turns ON, the 460L is
still operational, but a system fault or calibration fault exists that may make the current ozone
reading invalid.
The Invalid Reading output is turned ON for any of the following conditions:
The measured pressure in the measurement cell exceeds too high indicating that the O3
supply is under pressure.
The measured pressure in the measurement cell exceeds too low indicating that there
could be a blockage in the monitor’s internal pneumatic lines.
The measured concentration has exceeded the measurement range selected for the
monitor.
The measured concentration is an excessively negative reading (less than -0.10 ppm or –
10.0 ppb).
11.2.3. Lamp Low
The normal state for the lamp low output is OFF. This output turns on when the UV lamp intensity
as measured by the reference detector has dropped below 375mV. This could mean that the lamp
has drifted has started to drift low and that the lamp output should be adjusted.
11.2.4. Status LED / Status Output Troubleshooting Summary
Table 11-2 below is a logic truth table summarizing the recommended actions based on the states
of the four status front panel status LED’s. A ‘1’ indicates the LED is ON, a ‘0’ indicates the LED is
OFF, and ‘X’ indicates the LED can be in either state.
66
05228 Rev B
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
460L Instruction Manual
Troubleshooting
Table 11-2 Status LED / Output Trouble shooting Truth Table
SENSOR INVALID
LAMP
LOW
ACTIONS
OK
READING
Normal operation, no action required.
1
0
0
0
The Lamp intensity has drifted and is too high for proper operation.
Adjust the lamp.
0
0
The O3 sensor is not responding to request from the CPU for data.
UV Lamp output is starting to drift low.
1
0
X
X
1
1
Adjust the lamp output.
The UV lamp is either completely off or its intensity has drifted so low
that the monitor can not reliable measure O3.
Replace the UV lamp (see Section 9.3).
O3 reading is above the upper limit of the range selected for the
monitor.
Check or change the range setting (see Sections 7.3.9 to 7.3.11).
O3 reading is too negative.
Calibrate zero point (see Section 8.1).
X
1
X
Possible Internal blockage.
check or replace the filter (see Section 9.1.1).
O3 supply under high pressure.
Check gas lines from O3 source.
11.3. Concentration Alarm Outputs
Table 11-3 Alarm Output Troubleshooting
FAULT CONDITION
POSSIBLE CAUSES
Alarm mode is set for Non-Latching and Concentration value is ≥
relevant alarm set point.
Pressing ALARM ACKNOWLEDGE button does
not clear alarm or turn off Alarm Active LED.
ALARM ACKNOWLEDGE button is bad.
Using the ALMACK command over the serial
port does not clear alarm or turn off Alarm
Active LED.
Alarm mode is set for Non-Latching and Concentration value is ≥
relevant alarm set point.
LED has Failed.
Alarm Active LED doesn’t blink when one or
both Alarms are active
Display driver PCA malfunction
Alarm does not activate at expected O3
concentration
Set point for alarm is set incorrectly
Alarm Active Status Output OFF when HI alarm
or HI-HI Alarm is active
Electronic failure of Alarm Active status output.
Hi alarm or HI-HI alarm status output is OFF
when O3 concentration is ≥ related set point
value
Electronic failure of status output.
05228 Rev B
67
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
Troubleshooting
460L Instruction Manual
11.4. Technical Assistance
If this addendum and its trouble-shooting / repair sections do not solve your problems, technical
assistance may be obtained from:
Teledyne Instruments
Advanced Pollution Instrumentation Division
(TAPI)
Customer Service
9480 Carroll Park Drive
San Diego, California 92121-5201USA
Toll-free Phone: 800-324-5190
Phone: 858-657-9800
Fax: 858-657-9816
Email: [email protected]
Website: http://www.Teledyne-API.com
Before you contact Teledyne Instruments’ Customer service, fill out the problem report form in
Appendix C, which is also available online for electronic submission at http://www.Teledyne-
API.com/forms/p-fmapicom.asp.
USER NOTES:
68
05228 Rev B
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
A Primer on Electro-Static Discharge
460L Instruction Manual
12. A PRIMER ON ELECTRO-STATIC DISCHARGE
Teledyne Instruments considers the prevention of damage caused by the discharge of static
electricity to be extremely important part of making sure that your analyzer continues to provide
reliable service for a long time. This section describes how static electricity occurs, why it is so
dangerous to electronic components and assemblies as well as how to prevent that damage from
occurring.
12.1. How Static Charges are Created
Modern electronic devices such as the types used in the various electronic assemblies of your
analyzer, are very small, require very little power and operate very quickly. Unfortunately, the
same characteristics that allow them to do these things also make them very susceptible to
damage from the discharge of static electricity. Controlling electrostatic discharge begins with
understanding how electro-static charges occur in the first place.
Static electricity is the result of something called triboelectric charging which happens whenever
the atoms of the surface layers of two materials rub against each other. As the atoms of the two
surfaces move together and separate, some electrons from one surface are retained by the other.
Materials
Makes
Contact
Materials
Separate
+
+
+
+
PROTONS = 3
ELECTRONS = 2
PROTONS = 3
ELECTRONS = 4
PROTONS = 3
ELECTRONS = 3
PROTONS = 3
ELECTRONS = 3
NET CHARGE = -1
NET CHARGE = +1
NET CHARGE = 0
NET CHARGE = 0
Figure 12-1 Triboelectric Charging
If one of the surfaces is a poor conductor or even a good conductor that is not grounded, the
resulting positive or negative charge cannot bleed off and becomes trapped in place, or static.
The most common example of triboelectric charging happens when someone wearing leather or
rubber soled shoes walks across a nylon carpet or linoleum tiled floor. With each step, electrons
change places and the resulting electro-static charge builds up, quickly reaching significant levels.
Pushing an epoxy printed circuit board across a workbench, using a plastic handled screwdriver
or even the constant jostling of StyrofoamTM pellets during shipment can also build hefty static
charges.
05228 Rev B
DCN 5164
69
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
A Primer on Electro-Static Discharge
460L Instruction Manual
Table 12-1 Static Generation Voltages for Typical Activities
MEANS OF GENERATION
65-90% RH
1,500V
250V
10-25% RH
Walking across nylon carpet
Walking across vinyl tile
Worker at bench
35,000V
12,000V
6,000V
100V
Poly bag picked up from
bench
1,200V
1,500V
20,000V
18,000V
Moving around in a chair
padded with urethane foam
12.2. How Electro-Static Charges Cause Damage
Damage to components occurs when these static charges come into contact with an electronic
device. Current flows as the charge moves along the conductive circuitry of the device and the
typically very high voltage levels of the charge overheat the delicate traces of the integrated
circuits, melting them or even vaporizing parts of them. When examined by microscope the
damage caused by electro-static discharge looks a lot like tiny bomb craters littered across the
landscape of the component’s circuitry.
susceptibility levels, shows why Semiconductor Reliability News estimates that approximately
60% of device failures are the result of damage due to electro-static discharge.
Table 12-2 Sensitivity of Electronic Devices to Damage by ESD
DAMAGE SUSCEPTIBILITY
VOLTAGE RANGE
DEVICE
DAMAGE BEGINS
OCCURRING AT
CATASTROPHIC
DAMAGE AT
MOSFET
VMOS
10
100
30
1800
100
NMOS
60
GaAsFET
EPROM
60
2000
100
100
140
150
190
200
300
300
JFET
7000
500
SAW
Op-AMP
CMOS
2500
3000
2500
3000
Schottky Diodes
Film Resistors
This Film
Resistors
300
7000
ECL
500
500
500
500
SCR
1000
2500
Schottky TTL
Potentially damaging electro-static discharges can occur:
70
05228 Rev B
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
A Primer on Electro-Static Discharge
460L Instruction Manual
Any time a charged surface (including the human body) discharges to a device. Even
simple contact of a finger to the leads of a sensitive device or assembly can allow enough
discharge to cause damage. A similar discharge can occur from a charged conductive
object, such as a metallic tool or fixture.
When static charges accumulated on a sensitive device discharges from the device to
another surface such as packaging materials, work surfaces, machine surfaces or other
device. In some cases, charged device discharges can be the most destructive.
A typical example of this is the simple act of installing an electronic assembly into the
connector or wiring harness of the equipment in which it is to function. If the assembly is
carrying a static charge, as it is connected to ground a discharge will occur.
Whenever a sensitive device is moved into the field of an existing electro-static field, a
charge may be induced on the device in effect discharging the field onto the device. If the
device is then momentarily grounded while within the electrostatic field or removed from
the region of the electrostatic field and grounded somewhere else, a second discharge will
occur as the charge is transferred from the device to ground.
12.3. Common Myths About ESD Damage
I didn’t feel a shock so there was no electro-static discharge: The human nervous
system isn’t able to feel a static discharge of less than 3500 volts. Most devices are
damaged by discharge levels much lower than that.
I didn’t touch it so there was no electro-static discharge: Electro Static charges are
fields whose lines of force can extend several inches or sometimes even feet away from the
surface bearing the charge.
It still works so there was no damage: Sometimes the damaged caused by electro-static
discharge can completely sever a circuit trace causing the device to fail immediately. More
likely, the trace will be only partially occluded by the damage causing degraded
performance of the device or worse, weakening the trace. This weakened circuit may seem
to function fine for a short time, but even the very low voltage and current levels of the
device’s normal operating levels will eat away at the defect over time causing the device to
fail well before its designed lifetime is reached.
These latent failures are often the most costly since the failure of the equipment in which
the damaged device is installed causes down time, lost data, lost productivity, as well as
possible failure and damage to other pieces of equipment or property.
Static Charges can’t build up on a conductive surface: There are two errors in this
statement.
Conductive devices can build static charges if they are not grounded. The charge will be
equalized across the entire device, but without access to earth ground, they are still
trapped and can still build to high enough levels to cause damage when they are
discharged.
A charge can be induced onto the conductive surface and/or discharge triggered in the
presence of a charged field such as a large static charge clinging to the surface of a nylon
jacket of someone walking up to a workbench.
As long as my analyzer is properly installed, it is safe from damage caused by static
discharges: It is true that when properly installed the chassis ground of your analyzer is
tied to earth ground and its electronic components are prevented from building static
electric charges themselves. This does not prevent discharges from static fields built up on
05228 Rev B
DCN 5164
71
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
A Primer on Electro-Static Discharge
460L Instruction Manual
other things, like you and your clothing, from discharging through the instrument and
damaging it.
12.4. Basic Principles of Static Control
It is impossible to stop the creation of instantaneous static electric charges. It is not, however
difficult to prevent those charges from building to dangerous levels or prevent damage due to
electro-static discharge from occurring.
12.4.1. General Rules
Only handle or work on all electronic assemblies at a properly set up ESD station. Setting up an
ESD safe workstation need not be complicated. A protective mat properly tied to ground and a
wrist strap are all that is needed to create a basic anti-ESD workstation.
Protective Mat
Ground Point
Wrist Stra
Figure 12-2 Basic anti-ESD Workbench
For technicians that work in the field, special lightweight and portable anti-ESD kits are available
from most suppliers of ESD protection gear. These include everything needed to create a
temporary anti-ESD work area anywhere.
Always wear an Anti-ESD wrist strap when working on the electronic assemblies of
your analyzer. An anti-ESD wrist strap keeps the person wearing it at or near the same
potential as other grounded objects in the work area and allows static charges to dissipate
before they can build to dangerous levels. Anti-ESD wrist straps terminated with alligator
clips are available for use in work areas where there is no available grounded plug.
Also, anti-ESD wrist straps include a current limiting resistor (usually around one meg-ohm)
that protects you should you accidentally short yourself to the instrument’s power supply.
Simply touching a grounded piece of metal is insufficient. While this may temporarily
bleed off static charges present at the time, once you stop touching the grounded metal
new static charges will immediately begin to re-build. In some conditions, a charge large
enough to damage a component can rebuild in just a few seconds.
Always store sensitive components and assemblies in anti-ESD storage bags or bins:
Even when you are not working on them, store all devices and assemblies in a closed anti-
Static bag or bin. This will prevent induced charges from building up on the device or
assembly and nearby static fields from discharging through it.
Use metallic anti-ESD bags for storing and shipping ESD sensitive components and
assemblies rather than pink-poly bags. The famous, pink-poly bags are made of a
plastic that is impregnated with a liquid (similar to liquid laundry detergent) which very
slowly sweats onto the surface of the plastic creating a slightly conductive layer over the
surface of the bag.
72
05228 Rev B
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
A Primer on Electro-Static Discharge
460L Instruction Manual
While this layer may equalizes any charges that occur across the whole bag, it does not
prevent the build-up of static charges. If laying on a conductive, grounded surface, these
bags will allow charges to bleed away but the very charges that build up on the surface of
the bag itself can be transferred through the bag by induction onto the circuits of your ESD
sensitive device. Also, the liquid impregnating the plastic is eventually used up after which
the bag is as useless for preventing damage from ESD as any ordinary plastic bag.
Anti-Static bags made of plastic impregnated with metal (usually silvery in color) provide all
of the charge equalizing abilities of the pink-poly bags but also, when properly sealed,
create a Faraday cage that completely isolates the contents from discharges and the
inductive transfer of static charges.
Storage bins made of plastic impregnated with carbon (usually black in color) are also
excellent at dissipating static charges and isolating their contents from field effects and
discharges.
Never use ordinary plastic adhesive tape near an ESD sensitive device or to close an
anti-ESD bag. The act of pulling a piece of standard plastic adhesive tape, such as
Scotch® tape, from its roll will generate a static charge of several thousand or even tens of
thousands of volts on the tape itself and an associated field effect that can discharge
through or be induced upon items up to a foot away.
12.4.2. Basic anti-ESD Procedures for Analyzer Repair and
Maintenance
12.4.2.1. Working at the Instrument Rack
When working on the analyzer while it is in the instrument rack and plugged into a properly
grounded power supply
1. Attach you anti-ESD wrist strap to ground before doing anything else.
Use a wrist strap terminated with an alligator clip and attach it to a bare metal portion of
the instrument chassis.
This will safely connect you to the same ground level to which the instrument and all of its
components are connected.
2. Pause for a second or two to allow any static charges to bleed away.
3. Open the casing of the analyzer and begin work. Up to this point, the closed metal casing
of your analyzer has isolated the components and assemblies inside from any conducted
or induced static charges.
4. If you must remove a component from the instrument, do not lay it down on a non-ESD
preventative surface where static charges may lie in wait.
5. Only disconnect your wrist strap after you have finished work and closed the case of the
analyzer.
12.4.2.2. Working at an Anti-ESD Work Bench.
When working on an instrument of an electronic assembly while it is resting on a anti-ESD
workbench
1. Plug your anti-ESD wrist strap into the grounded receptacle of the work station before
touching any items on the work station and while standing at least a foot or so away.
05228 Rev B
73
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
A Primer on Electro-Static Discharge
460L Instruction Manual
This will allow any charges you are carrying to bleed away through the ground connection
of the workstation and prevent discharges due to field effects and induction from
occurring.
2. Pause for a second or two to allow any static charges to bleed away.
3. Only open any anti-ESD storage bins or bags containing sensitive devices or assemblies
after you have plugged your wrist strap into the workstation.
Lay the bag or bin on the workbench surface.
Before opening the container, wait several seconds for any static charges on the
outside surface of the container to be bled away by the workstation’s grounded
protective mat.
4. Do not pick up tools that may be carrying static charges while also touching or holding an
ESD sensitive Device.
5. Only lay tools or ESD-sensitive devices and assemblies on the conductive surface of your
workstation. Never lay them down on any non-ESD preventative surface.
6. Place any static sensitive devices or assemblies in anti-static storage bags or bins and
close the bag or bin before unplugging your wrist strap.
7. Disconnecting your wrist strap is always the last action taken before leaving the
workbench.
12.4.2.3. Transferring Components from Rack to Bench and Back
When transferring a sensitive device from an installed Teledyne Instruments analyzer to an Anti-
ESD workbench or back:
1. Follow the instructions listed above for working at the instrument rack and workstation.
2. Never carry the component or assembly without placing it in an anti-ESD bag or bin.
3. Before using the bag or container allow any surface charges on it to dissipate:
If you are at the instrument rack, hold the bag in one hand while your wrist strap is
connected to a ground point.
If you are at an anti-ESD workbench, lay the container down on the conductive work
surface.
In either case wait several seconds.
4. Place the item in the container.
74
05228 Rev B
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
A Primer on Electro-Static Discharge
460L Instruction Manual
5. Seal the container. If using a bag, fold the end over and fastening it with anti-ESD tape.
Folding the open end over isolates the component(s) inside from the effects of static
fields.
Leaving the bag open or simply stapling it shut without folding it closed prevents the
bag from forming a complete protective envelope around the device.
6. Once you have arrived at your destination, allow any surface charges that may have built
up on the bag or bin during travel to dissipate:
7. Connect your wrist strap to ground.
If you are at the instrument rack, hold the bag in one hand while your wrist strap is
connected to a ground point.
If you are at a anti-ESD workbench, lay the container down on the conductive work
surface.
In either case wait several seconds.
8. Open the container.
12.4.2.4. Opening Shipments from Teledyne Instruments’ Customer
Service.
Packing materials such as bubble pack and Styrofoam pellets are extremely efficient generators of
static electric charges. To prevent damage from ESD, Teledyne Instruments ships all electronic
components and assemblies in properly sealed ant-ESD containers.
Static charges will build up on the outer surface of the anti-ESD container during shipping as the
packing materials vibrate and rub against each other. To prevent these static charges from
damaging the components or assemblies being shipped make sure that you:
1. Always unpack shipments from Teledyne Instruments Customer Service by.
2. Open the outer shipping box away from the anti-ESD work area.
3. Carry the still sealed ant-ESD bag, tube or bin to the anti-ESD work area.
work station.
5. Reserve the anti-ESD container or bag to use when packing electronic components or
assemblies to be returned to Teledyne Instruments.
05228 Rev B
DCN 5164
75
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
A Primer on Electro-Static Discharge
460L Instruction Manual
12.4.2.5. Packing Components for Return to Teledyne Instruments
Customer Service
Always pack electronic components and assemblies to be sent to Teledyne Instruments Customer
Service in anti-ESD bins, tubes or bags.
CAUTION
ESD Hazard
DO NOT use pink-poly bags.
NEVER allow any standard plastic packaging materials to touch the
electronic component/assembly directly
This includes, but is not limited to, plastic bubble-pack, Styrofoam
peanuts, open cell foam, closed cell foam, and adhesive tape
DO NOT use standard adhesive tape as a sealer. Use ONLY anti-ESD
tape
1. Never carry the component or assembly without placing it in an anti-ESD bag or bin.
2. Before using the bag or container allow any surface charges on it to dissipate:
If you are at the instrument rack, hold the bag in one hand while your wrist strap is
connected to a ground point.
If you are at an anti-ESD workbench, lay the container down on the conductive work
surface.
In either case wait several seconds.
3. Place the item in the container.
76
05228 Rev B
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
A Primer on Electro-Static Discharge
460L Instruction Manual
4. Seal the container. If using a bag, fold the end over and fastening it with anti-ESD tape.
Folding the open end over isolates the component(s) inside from the effects of static
fields.
Leaving the bag open or simply stapling it shut without folding it closed prevents the
bag from forming a complete protective envelope around the device.
NOTE
If you do not already have an adequate supply of anti-ESD gags or containers available,
Teledyne Instruments’ Customer Service department will supply them
Follow the instructions listed above for working at the instrument rack and workstation.
05228 Rev B
DCN 5164
77
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
TELEDYNE INSTRUMENTS
460L Instruction Manual
Spare Parts List
USER NOTES:
78
05228 Rev B
DCN 5164
PRINTED DOCUMENTS ARE UNCONTROLLED
Download from Www.Somanuals.com. All Manuals Search And Download.
M460L Instruction Manual
Appendix A - Spare Parts list
Appendix A - Spare Parts list
•
05128 – Spare Parts List M460L
05229 Rev B
A-1
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix A - Spare Parts list
M460L Instruction Manual
A-2
05229 Rev B
Download from Www.Somanuals.com. All Manuals Search And Download.
M460L Spare Parts List
Part Number
025710100
Description
PCA, UV DETECTOR PREAMP, M454/M460
OVERLAY, FRONT PANEL, O3 SENSOR, M460L
PCA, FRONT PANEL, M454 PRODUCTS
PCA, EXPANSION BD, M554/M460X/03/CAL
ASSY, WATER DROP OUT FILTER, PFA, M450
AKIT, ELEMENTS, 47MM, 30 MICRON (10 PCS)
ASSY, UV LAMP (BIR) M452/454/460X/465L *
ASSY, FLOWMETER, 0-2.5L, M460M/L
CBL, PS AC IN, M454 NEMA
030500400
031680100
034920000
036280000
036750000
037420000
037490100
040530000
040540100
041980000
046170000
046740000
047110000
048490100
049680000
049910100
050120000
050170000
050300000
053010000
CN0000350
DR0000006
HW0000120
OP0000031
OR0000039
OR0000050
OR0000098
OR0000099
PS0000035
CBL, DC POWER, M460L NEMA
CBL ASSY, GROUND STRAP, 4.5"
ABSORPTION TUBE, M460L
ASSY, PUMP, 12VDC, M460M/M700E
ASSY, VALVE, M460L
PCA, O3 BENCH, M460L/M465L
ASSY, SENSOR, HI-CONC, M460L/M465L
PCA, MAINBOARD, M460L
MANUAL, OPERATORS, M460L
ASSY, DFU SAMPLE FILTER
ASSY, REF SCRUBBER, HI-CONC, M460L/M465L
AKIT, EXP KIT, CARULITE 200
CONNECTOR, 16 PIN, W/SCREW FLANGE
DRYER, 24", 1/4" SS FITTINGS
SHOCKMOUNT, GROMMET ISOLATOR
WINDOW, QUARTZ, 1/2"DIA, .063" THICK (KB
ORING, BENCH
ORING, BENCH
ORING, BENCH
ORING, WATER DROP OUT FILTER
EOS SWITCHING PS, 15V, 40W
05128C M460L- SPL.xls DCN5164
Page 1 of 1
02/18//09
Download from Www.Somanuals.com. All Manuals Search And Download.
THIS PAGE IS INTENTIONALLY LEFT BLANK
Download from Www.Somanuals.com. All Manuals Search And Download.
460L Instruction Manual
Appendix B – Electronic Schematics
Appendix B – Electronic Schematics
Table B-1
M460L List of Electronic Schematics
DOCUMENT #
03169
DESCRIPTION
PCA, FRONT PANEL DISPLAY
PCA, MAINBOARD
PCA, CPU
04992
03493
05230 Rev A
B-1
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix B – Electronic Schematics
460L Instruction Manual
B-2
05230 Rev A
Download from Www.Somanuals.com. All Manuals Search And Download.
|