		 Service Manual   
					DPO4000 Series   
					Digital Phosphor Oscilloscopes   
					071-1844-00   
					This document applies to firmware version 1.00   
					and above.   
					Warning   
					The servicing instructions are for use by qualified   
					personnel only. To avoid personal injury, do not   
					perform any servicing unless you are qualified to   
					do so. Refer to all safety summaries prior to   
					performing service.   
					
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Warranty 16   
					Tektronix warrants that the product will be free from defects in materials and workmanship for a period of three   
					(3) years from the date of original purchase from an authorized Tektronix distributor. If the product proves   
					defective during this warranty period, Tektronix, at its option, either will repair the defective product without   
					charge for parts and labor, or will provide a replacement in exchange for the defective product. Batteries are   
					excluded from this warranty. Parts, modules and replacement products used by Tektronix for warranty work may   
					be new or reconditioned to like new performance. All replaced parts, modules and products become the property   
					of Tektronix.   
					In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration   
					of the warranty period and make suitable arrangements for the performance of service. Customer shall be   
					responsible for packaging and shipping the defective product to the service center designated by Tektronix,   
					shipping charges prepaid, and with a copy of customer proof of purchase. Tektronix shall pay for the return of the   
					product to Customer if the shipment is to a location within the country in which the Tektronix service center is   
					located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for   
					products returned to any other locations.   
					This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate   
					maintenance and care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage   
					resulting from attempts by personnel other than Tektronix representatives to install, repair or service the product;   
					b) to repair damage resulting from improper use or connection to incompatible equipment; c) to repair any   
					damage or malfunction caused by the use of non-Tektronix supplies; or d) to service a product that has been   
					modified or integrated with other products when the effect of such modification or integration increases the time   
					or difficulty of servicing the product.   
					THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THE PRODUCT IN LIEU OF ANY   
					OTHER WARRANTIES, EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY   
					IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.   
					TEKTRONIX’ RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND   
					EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY.   
					TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL,   
					OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS   
					ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Download from Www.Somanuals.com. All Manuals Search And Download.   
				Table of Contents   
					General Safety Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   
					Service Safety Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   
					Environmental Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   
					iii   
					v 
					vii   
					Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   
					Manual Conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   
					Related Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   
					ix   
					ix   
					ix   
					Operating Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   
					1--1   
					Theory of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   
					Block Diagrams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   
					Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   
					I/O Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   
					Main Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   
					Front-Panel Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   
					2--1   
					2-1   
					2-2   
					2-2   
					2-2   
					2-3   
					Adjustment Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   
					Required Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   
					Overview of the Adjustment Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   
					Factory Adjustment Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   
					3--1   
					3-1   
					3-2   
					3-3   
					Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   
					Preventing ESD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   
					Inspection and Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   
					General Care . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   
					Interior Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   
					Exterior Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   
					Overview of Removal Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   
					Removal Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   
					Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   
					Troubleshooting Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   
					Unpacking and Repacking Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   
					4--1   
					4-1   
					4-1   
					4-2   
					4-2   
					4-2   
					4-6   
					4-8   
					4-12   
					4-13   
					4-16   
					Mechanical Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   
					Parts Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   
					Using the Replaceable Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   
					5--1   
					5-1   
					5-2   
					Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Index--1   
					i 
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Table of Contents   
					List of Tables   
					Table 4--1: External inspection check list . . . . . . . . . . . . . . . . . . . . . .   
					Table 4--2: Internal inspection check list . . . . . . . . . . . . . . . . . . . . . .   
					4--3   
					4--4   
					List of Figures   
					Figure 2--1: DPO4000 series block diagram . . . . . . . . . . . . . . . . . . . .   
					Figure 4--1: Locator for trim and cabinet removal . . . . . . . . . . . . . . .   
					Figure 4--2: Locator for internal modules . . . . . . . . . . . . . . . . . . . . . .   
					Figure 4--3: Primary troubleshooting procedure . . . . . . . . . . . . . . . .   
					Figure 4--4: AC power supply troubleshooting procedure . . . . . . . . .   
					Figure 4--5: Module isolation troubleshooting procedure . . . . . . . . .   
					Figure 5--1: Exploded view, case . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   
					2--1   
					4--7   
					4--8   
					4--13   
					4--14   
					4--15   
					5--4   
					Figure 5--2: Exploded view: front panel and display,   
					4-channel instruments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   
					5--6   
					5--8   
					Figure 5--3: Exploded view: front panel 2-channel instruments . . . .   
					Figure 5--4: Exploded view: power supply and I/O board . . . . . . . .   
					5--10   
					Figure 5--5: Exploded view: mainboard, inverter board,   
					and side fans . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   
					5--12   
					ii   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				General Safety Summary   
					Review the following safety precautions to avoid injury and prevent damage to   
					this product or any products connected to it.   
					To avoid potential hazards, use this product only as specified.   
					Only qualified personnel should perform service procedures.   
					Use Proper Power Cord. Use only the power cord specified for this product and   
					certified for the country of use.   
					To Avoid Fire or   
					Personal Injury   
					Connect and Disconnect Properly. Do not connect or disconnect probes or test   
					leads while they are connected to a voltage source.   
					Ground the Product. This product is grounded through the grounding conductor   
					of the power cord. To avoid electric shock, the grounding conductor must be   
					connected to earth ground. Before making connections to the input or output   
					terminals of the product, ensure that the product is properly grounded.   
					Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings   
					and markings on the product. Consult the product manual for further ratings   
					information before making connections to the product.   
					The inputs are not rated for connection to mains or Category II, III, or IV   
					circuits.   
					Connect the probe reference lead to earth ground only.   
					Do not apply a potential to any terminal, including the common terminal, that   
					exceeds the maximum rating of that terminal.   
					Do Not Operate Without Covers. Do not operate this product with covers or panels   
					removed.   
					Do Not Operate With Suspected Failures. If you suspect there is damage to this   
					product, have it inspected by qualified service personnel.   
					Avoid Exposed Circuitry. Do not touch exposed connections and components   
					when power is present.   
					Do Not Operate in Wet/Damp Conditions.   
					Do Not Operate in an Explosive Atmosphere.   
					Keep Product Surfaces Clean and Dry.   
					Provide Proper Ventilation. Refer to the manual’s installation instructions for   
					details on installing the product so it has proper ventilation.   
					iii   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				General Safety Summary   
					Terms in this Manual   
					These terms may appear in this manual:   
					WARNING. Warning statements identify conditions or practices that could result   
					in injury or loss of life.   
					CAUTION. Caution statements identify conditions or practices that could result in   
					damage to this product or other property.   
					Symbols and Terms   
					on the Product   
					These terms may appear on the product:   
					H 
					H 
					H 
					DANGER indicates an injury hazard immediately accessible as you read the   
					marking.   
					WARNING indicates an injury hazard not immediately accessible as you   
					read the marking.   
					CAUTION indicates a hazard to property including the product.   
					The following symbols may appear on the product:   
					Protective Ground   
					(Earth) Terminal   
					CAUTION   
					Refer to Manual   
					Mains Connected   
					ON (Power)   
					Push button OUT   
					Chassis Ground   
					Push button   
					IN   
					iv   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Service Safety Summary   
					Only qualified personnel should perform service procedures. Read this Service   
					Safety Summary and the General Safety Summary before performing any service   
					procedures.   
					Do Not Service Alone. Do not perform internal service or adjustments of this   
					product unless another person capable of rendering first aid and resuscitation is   
					present.   
					Disconnect Power. To avoid electric shock, switch off the instrument power, then   
					disconnect the power cord from the mains power.   
					Use Care When Servicing With Power On. Dangerous voltages or currents may   
					exist in this product. Disconnect power, remove battery (if applicable), and   
					disconnect test leads before removing protective panels, soldering, or replacing   
					components.   
					To avoid electric shock, do not touch exposed connections.   
					v 
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Service Safety Summary   
					vi   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Environmental Considerations   
					This section provides information about the environmental impact of the   
					product.   
					Observe the following guidelines when recycling an instrument or component:   
					Product End-of-Life   
					Handling   
					Equipment Recycling. Production of this equipment required the extraction and   
					use of natural resources. The equipment may contain substances that could be   
					harmful to the environment or human health if improperly handled at the   
					product’s end of life. In order to avoid release of such substances into the   
					environment and to reduce the use of natural resources, we encourage you to   
					recycle this product in an appropriate system that will ensure that most of the   
					materials are reused or recycled appropriately.   
					The symbol shown to the left indicates that this product   
					complies with the European Union’s requirements   
					according to Directive 2002/96/EC on waste electrical and   
					electronic equipment (WEEE). For information about   
					recycling options, check the Support/Service section of the   
					Tektronix Web site (www.tektronix.com).   
					Mercury Notification. This product uses an LCD backlight lamp that contains   
					mercury. Disposal may be regulated due to environmental considerations. Please   
					contact your local authorities or, within the United States, the Electronics   
					
					This product has been classified as Monitoring and Control equipment, and is   
					outside the scope of the 2002/95/EC RoHS Directive. This product is known to   
					contain lead, cadmium, mercury, and hexavalent chromium.   
					Restriction of Hazardous   
					Substances   
					vii   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Environmental Considerations   
					viii   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Preface   
					This Service manual provides information to troubleshoot, disassemble, and   
					replace parts on the DPO4000 Series Digital Phosphor Oscilloscopes.   
					Manual Conventions   
					This manual uses certain conventions that you should become familiar with   
					before attempting service.   
					Modules   
					Throughout this manual, any replaceable component, assembly, or part is   
					referred to by the term module. A module is composed of electrical and   
					mechanical assemblies, circuit cards, interconnecting cables, and user-accessible   
					controls.   
					Replaceable Parts   
					This manual refers to any field-replaceable assembly or mechanical part   
					specifically by its name or generically as a replaceable part. In general, a   
					replaceable part is any circuit board or assembly, such as the hard disk drive, or a   
					mechanical part, such as the I/O port connectors, that is listed in the replaceable   
					parts list.   
					Safety   
					Symbols and terms related to safety appear in the Service Safety Summary found   
					at the beginning of this manual.   
					Related Documentation   
					To read about   
					Use these documents   
					Installation and Operation   
					DPO4000 Series Digital Phosphor Oscilloscopes User   
					Manual (available in 11 languages)   
					Specifications and Performance DPO4000 Technical Reference (PDF only)   
					Programmer Commands DPO4000 Programmer Manual (PDF only)   
					Analysis and Connectivity Tools Getting Started with OpenChoicet Solutions Manual   
					Installing and testing applica-   
					tions modules   
					DPO4000 Series Application Modules Installation Instructions   
					Manual   
					ix   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Preface   
					x 
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Operating Information   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Download from Www.Somanuals.com. All Manuals Search And Download.   
				Operating Information   
					For information on installing and operating your DPO4000 Series Digital   
					Phosphor Oscilloscope, please refer to the DPO4000 Series Digital Phosphor   
					Oscilloscopes User Manual. This manual is available on the Web at   
					
					1-1   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Operating Information   
					1-2   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Theory of Operation   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Download from Www.Somanuals.com. All Manuals Search And Download.   
				Theory of Operation   
					This chapter describes the electrical operation of the oscilloscope to the module   
					level. Figure 2--1 shows the oscilloscope module interconnections.   
					Block Diagrams   
					Display   
					back light   
					inverter   
					power   
					LCD   
					Display   
					Aux Out   
					J170   
					J930   
					J680   
					AC Line   
					Power   
					supply   
					Front panel   
					J1   
					J100   
					I/O Board   
					J440   
					Input ports for   
					application   
					modules   
					J0480   
					J610   
					Compact   
					Flash Port   
					J420, J421   
					Main board   
					J120   
					J130   
					J570   
					J571   
					J850   
					USB   
					J510   
					Bezel   
					buttons   
					3 Fans   
					Attenuator circuit   
					Ethernet   
					connector   
					Ch 1   
					Ch 1   
					Ch 2   
					Ch 2   
					Ch 3   
					Ch 4   
					EXT   
					EXT 4-channel oscilloscopes   
					2-channel oscilloscopes   
					Figure 2-1: DPO4000 series block diagram   
					2-1   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Theory of Operation   
					Power Supply   
					The power supply converts AC line voltage to +12V to power all internal   
					circuits.   
					I/O Board   
					The I/O board contains USB ports, Compact Flash port, an Ethernet connector,   
					TekLink, XGA Video, and a Trigger Out port.   
					Main Board   
					The main board module contains the following functions:   
					Acquisition System   
					The acquisition system begins with the analog signal path and ends with a   
					digitized signal in memory. The signal enters a channel input, and then passes   
					through an attenuator and preamplifier. The analog signal from each preamplifier   
					goes through a digitizer, a time base controller, and then into acquisition   
					memory. The analog signal from each preamplifier is also distributed to a trigger   
					circuit.   
					Trigger System   
					Display System   
					The trigger system processes the analog signals from the preamplifiers and sends   
					trigger information to the time base controller. Advanced trigger functions are   
					enabled only when the appropriate application modules and supporting software   
					are installed.   
					The display system combines live waveform data from acquisition memory with   
					menus and text and stores this information in display memory. It then uses this   
					data to refresh the XGA display module (LCD).   
					2-2   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Theory of Operation   
					Processor System   
					The processor system contains a 440EP Power PC microprocessor that controls   
					the entire instrument. The processor system also contains FLASH ROM, system   
					RAM, and interfaces to USB ports and a Compact flash slot.   
					Power Converter   
					The power converter receives +12V from the main power supply and generates   
					voltages for the analog and digital circuitry on the main board.   
					Front-Panel Board   
					The front-panel board contains a microprocessor that reads the front-panel   
					buttons and controls, and then sends this information to the processor system on   
					the main board. The front-panel board also generates the probe compensation   
					output signal and provides an interface to the application modules.   
					2-3   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Theory of Operation   
					2-4   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Adjustment Procedures   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Download from Www.Somanuals.com. All Manuals Search And Download.   
				Adjustment Procedures   
					This chapter contains the factory adjustment procedures for the DPO4000 series   
					oscilloscopes. Only qualified personnel should perform adjustment procedures.   
					Read the Service Safety Summary and the General Safety Summary before   
					performing any service procedures.   
					NOTE. The voltage references inside the oscilloscope are very stable over time   
					and should not require routine adjustment. Before performing any procedure in   
					this chapter, do the Performance Verification procedures to check whether the   
					oscilloscope meets specifications. See the DPO4000 Technical Reference.   
					Successful completion of this adjustment procedure automatically updates the   
					instrument Calibration Due date and time. (Completion of the Performance   
					Verification procedure does not update the instrument Calibration Due date and   
					time.)   
					Required Equipment   
					The following equipment, or a suitable equivalent, is required to complete these   
					procedures.   
					Description   
					Minimum requirements   
					Example   
					DC voltage source   
					50 mV to 70 V, 0.1% amplitude Wavetek 9500 Oscilloscope   
					accuracy   
					Calibrator with one 9510   
					Output Module   
					Sine Generator   
					Edge Generator   
					50 Ω BNC cable   
					1 GHz   
					Wavetek 9500 with 9530   
					output modules   
					1 kHz with <50 ps ch-ch skew   
					BNC male to BNC male, ≈ 10 in Tektronix part number   
					(25 cm) long 012-0208-00   
					Adjustment tool   
					Non conducting shaft with diame- Tektronix part number   
					ter ≤2.5 mm (≤0.1 in) 003-1433-00   
					3-1   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Adjustment Procedures   
					Overview of the Adjustment Process   
					Before performing adjustment procedures, you must warm up the oscilloscope at   
					least ten minutes in an ambient temperature between 20 °C and 30 °C. Adjust-   
					ments performed before warm-up or outside this temperature range may result in   
					poor performance.   
					The factory adjustment procedure consists of a series of steps; as you move   
					through these steps, the oscilloscope display provides instructions that describe   
					the specific input signal requirements for each step. If the the oscilloscope passes   
					the step, it moves on to the next step. If the oscilloscope fails, you can repeat the   
					step or choose to abort the procedure.   
					NOTE. Do not turn any knobs or push any front-panel buttons other than the   
					Next Step or Previous Step buttons during the adjustment procedure. Doing so   
					will cause the oscilloscope to abort the adjustment procedure. The oscilloscope   
					uses the previous calibration constants if the adjustment procedure is aborted.   
					The screen does not display the actual oscilloscope settings (such as channel   
					input impedance or vertical and horizontal settings) during the adjustment   
					procedure. The oscilloscope automatically sets the instrument settings, but these   
					settings may not read out correctly on the display.   
					During some steps, the instrument may appear to be idle for several minutes   
					while it is processing information internally.   
					If the oscilloscope completes all steps in the procedure successfully, a “Pass”   
					message is displayed and the new calibration constants take effect. If the   
					oscilloscope does not pass the procedure, the oscilloscope reverts to the previous   
					calibration settings.   
					3-2   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Adjustment Procedures   
					Factory Adjustment Procedure   
					To perform the factory adjustment procedure, do these steps:   
					1. Connect the oscilloscope to an AC power source.   
					NOTE. You must connect the oscilloscope and the test equipment to the same AC   
					power circuit. Connect the oscilloscope and test instruments to a common power   
					strip if you are unsure of the AC power circuit distribution. Connecting the   
					oscilloscope and test instruments to separate AC power circuits can result in   
					offset voltages between the equipment, which can invalidate the adjustment   
					procedure.   
					2. Locate the hole in the rear panel that provides access to the CAL switch on   
					the rear panel.   
					CAL   
					3. Insert the adjustment tool into the CAL switch hole. Use the tool to push and   
					hold the CAL switch in while you power on the oscilloscope. Release the   
					switch when the oscilloscope shows the power-on message with the   
					oscilloscope graticule in the background.   
					4. Push the Default Setup button.   
					NOTE. Do not skip the recall Default Setup step. You must recall the factory   
					settings in order to correctly start the factory adjustment procedure.   
					5. Push the MENU OFF button.   
					6. Push the UTILITY button.   
					7. Push the System Config lower bezel button repeatedly to select Calibra-   
					tion.   
					3-3   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Adjustment Procedures   
					8. Push the Factory lower bezel button.   
					9. Push the OK Do Factory Calibration side bezel button to start the   
					adjustment process.   
					10. Connect signals from the DC voltage source as instructed by the oscilloscope   
					display. After you connect the correct signal, push the OK Do Next Step   
					side bezel button to continue the process.   
					NOTE. The oscilloscope adjusts itself automatically using the reference signal   
					from the DC voltage source. You do not need to make any adjustments.   
					If you make an error, such as connecting the wrong input signal, you can repeat   
					the last step by pushing the Return to Previous Screen side bezel button.   
					11. Continue with the adjustment process until it is complete. You can abort the   
					process at any time by pushing the MENU OFF button.   
					12. Do the performance verification tests to verify that the adjustment procedure   
					has correctly calibrated the oscilloscope. (See the DPO4000 Technical   
					Reference.) Incorrect use of calibration equipment can cause the oscilloscope   
					to pass the adjustment procedure but fail performance verification.   
					3-4   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Maintenance   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Download from Www.Somanuals.com. All Manuals Search And Download.   
				Maintenance   
					This section contains the information needed to do periodic and corrective   
					maintenance on the oscilloscope as well as repackaging instructions if you need   
					to return the oscilloscope to Tektronix for service.   
					Preventing ESD   
					Before servicing this product, read the Safety Summary and Introduction at the   
					front of the manual and the electrostatic discharge (ESD) information below.   
					CAUTION. Static discharge can damage any semiconductor component in this   
					oscilloscope.   
					When performing any service that requires internal access to the oscilloscope,   
					adhere to the following precautions to avoid damaging internal modules and their   
					components due to electrostatic discharge.   
					1. Minimize handling of static-sensitive circuit boards and components.   
					2. Transport and store static-sensitive modules in their static protected   
					containers or on a metal rail. Label any package that contains static-sensitive   
					boards.   
					3. Discharge the static voltage from your body by wearing a grounded antistatic   
					wrist strap while handling these modules. Do service of static-sensitive   
					modules only at a static-free work station.   
					4. Do not place anything capable of generating or holding a static charge on the   
					work station surface.   
					5. Handle circuit boards by the edges when possible.   
					6. Do not slide the circuit boards over any surface.   
					7. Avoid handling circuit boards in areas that have a floor or work-surface   
					covering capable of generating a static charge.   
					Inspection and Cleaning   
					Inspection and cleaning are done as preventive maintenance. Preventive   
					maintenance, when done regularly, may prevent oscilloscope malfunction and   
					enhance its reliability.   
					4-1   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Maintenance   
					Preventive maintenance consists of visually inspecting and cleaning the   
					oscilloscope and using general care when operating it.   
					How often you do maintenance depends on the severity of the environment in   
					which the oscilloscope is used. A proper time to perform preventive maintenance   
					is just before oscilloscope adjustment.   
					General Care   
					The cabinet helps keep dust out of the oscilloscope and should normally be in   
					place when operating the oscilloscope.   
					Flat Panel Display   
					Cleaning   
					The display is a soft plastic display and must be treated with care during   
					cleaning.   
					CAUTION. Improper cleaning agents or methods can damage the flat panel   
					display.   
					Avoid using abrasive cleaners or commercial glass cleaners to clean the display   
					surface.   
					Avoid spraying liquids directly on the display surface.   
					Avoid scrubbing the display with excessive force.   
					Clean the flat panel display surface by gently rubbing the display with a   
					clean-room wipe (such as Wypall Medium Duty Wipes, #05701, available from   
					Kimberly-Clark Corporation).   
					Interior Cleaning   
					Use a dry, low-velocity stream of air to clean the interior of the chassis. Use a   
					soft-bristle, non-static-producing brush for cleaning around components. If you   
					must use a liquid for minor interior cleaning, use a 75% isopropyl alcohol   
					solution and rinse with deionized water.   
					WARNING. To avoid injury, power down the instrument and disconnect it from   
					line voltage before performing any procedure that follows.   
					Exterior Cleaning   
					Clean the exterior surfaces of the chassis with a dry lint-free cloth or a soft-   
					bristle brush. If any dirt remains, use a cloth or swab dipped in a 75% isopropyl   
					alcohol solution. Use a swab to clean narrow spaces around controls and   
					connectors. Do not use abrasive compounds on any part of the chassis that may   
					damage the chassis.   
					Clean the On/Standby switch using a dampened cleaning towel. Do not spray or   
					wet the switch directly.   
					4-2   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Maintenance   
					CAUTION. Avoid the use of chemical cleaning agents, which might damage the   
					plastics used in this oscilloscope. Use only deionized water when cleaning the   
					menu buttons or front-panel buttons. Use a 75% isopropyl alcohol solution as a   
					cleaner and rinse with deionized water. Before using any other type of cleaner,   
					consult your Tektronix Service Center or representative.   
					Lubrication. There is no periodic lubrication required for this oscilloscope.   
					Exterior Inspection   
					Inspect the outside of the oscilloscope for damage, wear, and missing parts,   
					using Table 4--1 as a guide. Immediately repair defects that could cause personal   
					injury or lead to further damage to the oscilloscope.   
					Table 4-1: External inspection check list   
					Item   
					Inspect for   
					Repair action   
					Cabinet, front panel,   
					and cover   
					Cracks, scratches, deformations, Repair or replace defective   
					damaged hardware   
					module   
					Front-panel knobs   
					Missing, damaged, or loose   
					knobs   
					Repair or replace missing or   
					defective knobs   
					Connectors   
					Broken shells, cracked insulation, Repair or replace defective   
					and deformed contacts. Dirt in   
					connectors   
					modules. Clear or wash out dirt.   
					Carrying handle, and   
					cabinet feet   
					Correct operation   
					Repair or replace defective   
					module   
					Accessories   
					Missing items or parts of items,   
					bent pins, broken or frayed   
					Repair or replace damaged or   
					missing items, frayed cables, and   
					cables, and damaged connectors defective modules   
					If the display is very dirty, moisten the wipe with distilled water or a 75%   
					isopropyl alcohol solution and gently rub the display surface. Avoid using excess   
					force or you may damage the plastic display surface.   
					CAUTION. To prevent getting moisture inside the oscilloscope during external   
					cleaning, use only enough liquid to dampen the cloth or applicator.   
					4-3   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Maintenance   
					Interior Inspection   
					To access the inside of the oscilloscope for inspection and cleaning, refer to the   
					Removal Procedures in this section.   
					Inspect the internal portions of the oscilloscope for damage and wear, using   
					Table 4--2 as a guide. Repair any defects immediately.   
					If any circuit board is repaired or replaced, check Table 4--2 to see if it is   
					necessary to adjust the oscilloscope.   
					CAUTION. To prevent damage from electrical arcing, ensure that circuit boards   
					and components are dry before applying power to the oscilloscope.   
					Table 4-2: Internal inspection check list   
					Item   
					Inspect for   
					Repair action   
					Circuit boards   
					Loose, broken, or corroded   
					solder connections. Burned   
					circuit boards. Burned, broken, or   
					cracked circuit-run plating.   
					Remove and replace damaged   
					circuit board.   
					Resistors   
					Burned, cracked, broken, blis-   
					tered condition.   
					Remove and replace damaged   
					circuit board.   
					Solder connections   
					Capacitors   
					Cold solder or rosin joints.   
					Resolder joint and clean with   
					isopropyl alcohol.   
					Damaged or leaking cases.   
					Corroded solder on leads or   
					terminals.   
					Remove and replace damaged   
					circuit board.   
					Wiring and cables   
					Chassis   
					Loose plugs or connectors.   
					Burned, broken, or frayed wiring. replace modules with defective   
					wires or cables.   
					Firmly seat connectors. Repair or   
					Dents, deformations, and dam-   
					aged hardware.   
					Straighten, repair, or replace   
					defective hardware.   
					4-4   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Maintenance   
					Cleaning Procedure — Interior. To clean the oscilloscope interior, do the following   
					steps:   
					1. Blow off dust with dry, low-pressure, deionized air (approximately 9 psi).   
					2. Remove any remaining dust with a lint-free cloth dampened in isopropyl   
					alcohol (75% solution) and rinsed with warm deionized water. (A cotton-   
					tipped applicator is useful for cleaning in narrow spaces and on circuit   
					boards.)   
					STOP. If, after doing steps 1 and 2, a module is clean upon inspection, skip the   
					following steps.   
					If steps 1 and 2 do not remove all the dust or dirt, the oscilloscope may be spray   
					washed using a solution of 75% isopropyl alcohol by doing the following steps:   
					1. Gain access to the parts to be cleaned by removing easily accessible shields   
					and panels (see Removal Procedures).   
					2. Spray wash dirty parts with the isopropyl alcohol and wait 60 seconds for the   
					majority of the alcohol to evaporate.   
					3. Use hot (120 _F to 140 _F) deionized water to thoroughly rinse them.   
					4. Dry all parts with low-pressure, deionized air.   
					5. Dry all components and assemblies in an oven or drying compartment using   
					low-temperature (125 _F to 150 _F) circulating air.   
					4-5   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Maintenance   
					Overview of Removal Procedures   
					Refer to Figures 4--1 and 4--2 for the location of modules that you are removing   
					or installing. If you are disassembling the instrument for cleaning, refer to the   
					Inspection and Cleaning procedure on page 4--1 for instructions.   
					List of Modules   
					The Mechanical Parts List chapter provides a list of all replaceable modules.   
					Any replaceable component, assembly, or part is referred to by the term module.   
					WARNING. Only qualified personnel should perform service procedures. Before   
					doing this or any other procedure in this manual, read the General Safety   
					Summary and Service Safety Summary located at the beginning of this manual.   
					Also, to prevent possible injury to service personnel or damage to electrical   
					components, read Preventing ESD on page 4--1.   
					Tools Required   
					You will need the following tools to remove and replace all the modules in the   
					instrument.   
					H 
					Torque-limiting screwdriver, 12 in ⋅ lb (1.3 N ⋅ m) range with TORX T-15   
					tips. (If you are using a magnetic screwdriver with interchangeable bits, the   
					T-15 bit must be a long bit, at least 5 cm (2 in) overall length.)   
					H 
					H 
					H 
					H 
					Screwdriver with 1   
					8 inch flat blade   
					@ 
					Phillips screwdrivers with P0 and P2 tips   
					3 
					1@ 6 inch nut driver   
					Needle-nose pliers   
					Use the following tables to remove the trim, cabinet, and internal modules.   
					Trim, Cabinet, and Module   
					Removal   
					WARNING. Before doing any procedure in this subsection, disconnect the power   
					cord from the line voltage source. Failure to do so could cause serious injury or   
					death.   
					4-6   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Maintenance   
					Rear case   
					Front foot   
					Front cover   
					Figure 4-1: Locator for trim and cabinet removal   
					4-7   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Maintenance   
					Power supply   
					assembly   
					Inverter board   
					assembly   
					Side fans   
					assembly   
					Display   
					assembly   
					I/O board   
					assembly   
					Main board   
					Front Case   
					assembly   
					Front Panel   
					assembly   
					Figure 4-2: Locator for internal modules   
					Removal Procedures   
					These procedures assume you have access to the module that you are removing.   
					Refer to the exploded drawings on pages 5--4 through 5--12 and to Figures 4--1   
					and 4--2, on pages 4--7 and 4--8 to determine which trim and or modules to   
					remove.   
					CAUTION. When removing or installing the keypad, make sure you do not touch   
					the switch contacts with your fingers. The oils in your fingers will degrade or   
					damage the switch contacts. To help prevent damage to the keypad use cotton   
					gloves when removing or installing the keyboard pad.   
					4-8   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Maintenance   
					CAUTION. To avoid damage to the front panel Standby/On switch assembly, do   
					not set the Display module assembly on a work surface. Sliding the oscilloscope   
					over the edge of the work surface could break off the On/Standby switch   
					assembly.   
					Rear Case   
					Follow these steps to remove the rear case. Use a TORX T-15 screwdriver tip.   
					See Figure 5--1 on page 5--4.   
					1. Remove the 4 screws from the back and sides of the rear case.   
					2. Grasp the case and pull outwards to remove it.   
					Power Supply Module   
					Follow these steps to remove the power supply module. You need to have   
					previously removed the rear case. Use a TORX T-15 screwdriver tip. See   
					Figure 5--4 on page 5--10.   
					1. Remove the 7 screws that secure the power supply module to the chassis.   
					2. Disconnect the power supply fan and connectors from the I/O board and the   
					main board.   
					3. Lift the power supply module off the chassis.   
					Power Supply Fan. Follow these steps to remove the power supply fan assembly.   
					You need to have previously removed the rear case and power supply assembly.   
					Use a TORX T-15 and a P2 screwdriver tip. See Figure 5--4 on page 5--10.   
					1. Remove the 5 screws that attach the power supply cover.   
					2. Lift the power supply cover off.   
					3. Use the P2 screwdriver tip and remove the 2 screws that attach the fan to the   
					cover.   
					Power Supply. Follow these steps to remove the power supply. You need to have   
					previously removed the rear case and power supply assembly. Use a TORX T-15   
					screwdriver tip. See Figure 5--4 on page 5--10.   
					1. Remove the 5 screws that attach the power supply cover.   
					2. Lift the power supply cover off.   
					3. Remove the 4 screws that attach the power supply to the power supply   
					bracket.   
					4-9   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Maintenance   
					I/O Board   
					Follow these steps to remove the I/O board. You need to have previously   
					removed the rear case and power supply module. Use TORX T-15 screwdriver   
					tip. See Figure 5--4 on page 5--10.   
					1. Remove the 5 screws that hold the I/O board to the chassis.   
					2. Lift straight up to remove.   
					Main Fans   
					Follow these steps to remove the side fan assembly. You need to have previously   
					removed the rear case, power supply assembly, and I/O board. Use a TORX T-15   
					screwdriver tip. See Figure 5--5 on page 5--12.   
					1. Remove the 6 screws that hold the side fans to the chassis.   
					2. Disconnect the two fan connectors from the main board.   
					3. Slide the fan assembly out and away from the chassis.   
					Main Board   
					Follow these steps to remove the main board. You need to have previously   
					removed the rear case, power supply, I/O board, and side fans. Use a TORX T-15   
					screwdriver tip. See Figure 5--5 on page 5--12.   
					1. Disconnect the display cable from the main board.   
					2. Disconnect the backlight cable from the inverter board.   
					3. Remove 13 screws from the main board that connect it to the chassis.   
					4. Lift the main board up to remove.   
					Inverter board. Follow these steps to remove the inverter board from the main   
					board. You must have previously removed the main board. Use a TORX T-15   
					screwdriver tip. See Figure 5--5 on page 5--12.   
					1. Disconnect the inverter board cable from the main board.   
					2. Remove the inverter board from the plastic bracket.   
					3. Remove the 2 screws that hold the inverter board bracket to the main board.   
					4. Remove the inverter board bracket from the main board.   
					4-10   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Maintenance   
					Front Case Assembly   
					Follow these steps to remove the front case assembly. You must have previously   
					removed the Rear Case Assembly. See Figure 5--2 on page 5--6.   
					1. Pull off 14 knobs (2 large, 5 medium, 6 small, and 1 rotary).   
					2. Remove the front case by pulling back the board snaps and lifting the lower   
					right corner of the front case.   
					3. Place the oscilloscope face up on a soft surface (such as an anti-static mat),   
					with the bottom facing you.   
					4. Disconnect the bezel button flex circuit by gently pulling the flex circuit out   
					of the connector.   
					Display   
					Follow these steps to remove the display module. You need to have previously   
					removed the rear case, power supply assembly, I/O board, and front case   
					assembly. Use a TORX T-15 screwdriver tip. See Figure 5--2 on page 5--6.   
					1. Disconnect the display power cable from the main board and the backlight   
					cable from the inverter board.   
					2. Remove the 4 screws that hold the display to the front chassis.   
					3. Gently lift the display up and off of the chassis.   
					CAUTION. Be careful when removing and reinstalling the Display module cables.   
					If the connectors have bent pins or are installed incorrectly; the Display may be   
					destroyed.   
					Front Panel   
					Follow these steps to remove the front panel module. You need to have   
					previously removed the front case assembly. Use a TORX T-15 screwdriver tip.   
					See Figure 5--2 on page 5--6.   
					1. Remove the 9 screws holding the front panel board to the chassis.   
					2. Lift the front panel board off of the chassis.   
					4-11   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Maintenance   
					Troubleshooting   
					This section contains information and procedures to help you isolate a defective   
					module in a DPO4000 series oscilloscope.   
					WARNING. Before doing this or any other procedure in this manual, read the   
					General Safety Summary and Service Safety Summary found at the beginning of   
					this manual. Also, to prevent possible injury to service personnel or damage to   
					electrical components, read Preventing ESD on page 4--1.   
					Adjustment After Repair   
					If you replace the main board, you must adjust the instrument after repair. Refer   
					Adjustment Procedures beginning on page 3--1 for information about adjustment.   
					Required Tools and   
					Equipment   
					You need the following equipment to troubleshoot the instrument.   
					Tools and Equipment   
					Example   
					DMM   
					3.5 or greater digit DMM   
					4-12   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Maintenance   
					Troubleshooting Procedure   
					Figures 4--3 through 4--5 are troubleshooting procedure flowcharts. Use them to   
					troubleshoot an instrument failure. Begin with Figure 4--3.   
					NOTE. Before performing the troubleshooting procedure, remove the instrument   
					cabinet (see page 4--9 ).   
					Power on the   
					instrument.   
					Does   
					the instrument   
					display the power-up   
					screen?   
					Perform the AC Power Supply   
					troubleshooting procedure   
					(Figure 4-4).   
					No   
					Yes   
					Does   
					the instrument pass   
					all the   
					Perform the Module Isolation   
					troubleshooting procedure   
					(Figure 4-5).   
					No   
					diagnostics?   
					Yes   
					Done.   
					Figure 4-3: Primary troubleshooting procedure   
					4-13   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Maintenance   
					Continued from   
					Figure 4-3.   
					¯ Power off oscilloscope.   
					¯ Disconnect AC power cord.   
					¯ Remove instrument case.   
					¯ Check that all cables are   
					connected and fully seated into   
					board connectors.   
					No   
					¯ Reconnect loose cables.   
					¯ Return to top of Figure 4-3.   
					Are all cables   
					connected?   
					Yes   
					Connect AC power cord.   
					¯ Measure for +12 V DC ( 0.3   
					V) on the connector pins. (See   
					figure at right)   
					!WARNING!   
					Be careful when measuring near   
					the AC power supply, as there   
					are line voltages present on   
					exposed parts.   
					Are there   
					+12 volts on the   
					No   
					Replace the AC Power   
					Supply module.   
					connector   
					pins ?   
					Yes   
					AC power supply OK. Return   
					the DPO4000 oscilloscope to   
					the nearest Tektronix Service   
					Center for repair.   
					Figure 4-4: AC power supply troubleshooting procedure   
					4-14   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Maintenance   
					From the oscilloscope   
					power-up diagnostics   
					message display.   
					Does   
					the power-up   
					screen report any   
					errors?   
					No   
					Oscilloscope is operating OK.   
					Yes   
					¯ Push UTILITY.   
					¯ Push System to select   
					Self Test.   
					¯ Push OK Run Self Test.   
					¯ Determine failure from   
					Self Test Results.   
					¯ Use table to select module to   
					replace.   
					Failed Test   
					Most Likely Cause   
					Acquisition   
					CPU   
					Display   
					Front Panel   
					I/O   
					Main board   
					Main board   
					Main board   
					Front Panel (or Main board)   
					Comm module (or Main board)   
					Main board   
					ROM   
					App Modules   
					Front Panel (or App Keys)   
					Figure 4-5: Module isolation troubleshooting procedure   
					4-15   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Maintenance   
					Unpacking and Repacking Instructions   
					This section contains the information needed to unpack the oscilloscope and   
					repack it for shipment or storage.   
					Unpacking   
					The oscilloscope and its standard accessories are carefully packed at the factory   
					in a shipping carton. If, upon receipt, damage to the shipping carton is evident,   
					notify the shipper. Tektronix, Inc. is not responsible for damage caused during   
					shipping.   
					If you have not already done so, carefully remove the oscilloscope and its   
					accessories from the shipping carton and inspect them for damage. Save the   
					shipping carton for repacking or storage.   
					Repacking   
					Use a corrugated cardboard shipping carton having a test strength of at least   
					275 pounds (125 kg) and with an inside dimension at least six inches (15.25 cm)   
					greater than the instrument dimensions.   
					If the instrument is being shipped to a Tektronix Service Center, enclose the   
					following information:   
					H 
					H 
					H 
					H 
					H 
					The owner’s address   
					Name and phone number of a contact person   
					Type and serial number of the instrument   
					Reason for returning   
					A complete description of the service required   
					Seal the shipping carton with an industrial stapler or strapping tape.   
					Mark the address of the Tektronix Service Center and also your own return   
					address on the shipping carton in two prominent locations.   
					The oscilloscope should be stored in a clean, dry environment. The following   
					environmental characteristics apply for both shipping and storage:   
					Storage   
					H 
					H 
					Temperature range: --20 °C to +60 °C   
					Altitude: To 15,000 m   
					See the DPO4000 Technical Reference for a complete listing of the environmen-   
					tal characteristics.   
					4-16   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Mechanical Parts List   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Download from Www.Somanuals.com. All Manuals Search And Download.   
				Mechanical Parts List   
					This chapter contains a list of the replaceable modules for the DPO4000 series   
					oscilloscope. Use this list to identify and order replacement parts.   
					Parts Ordering Information   
					Replacement parts are available through your local Tektronix field office or   
					representative.   
					Changes to Tektronix products are sometimes made to accommodate improved   
					components as they become available and to give you the benefit of the latest   
					improvements. Therefore, when ordering parts, it is important to include the   
					following information in your order.   
					H 
					H 
					H 
					H 
					Part number (see Part Number Revision Level below)   
					Instrument type or model number   
					Instrument serial number   
					Instrument modification number, if applicable   
					If you order a part that has been replaced with a different or improved part, your   
					local Tektronix field office or representative will contact you concerning any   
					change in part number.   
					Change information, if any, is located at the rear of this manual.   
					Tektronix part numbers contain two digits that show the revision level of the   
					part. For some parts in this manual, you will find the letters XX in place of the   
					revision level number.   
					Part Number Revision   
					Level   
					Part Number Revision Level   
					Revision Level May Show as XX   
					670-7918-03   
					670-7918-XX   
					When you order parts, Tektronix will provide you with the most current part for   
					your product type, serial number, and modification (if applicable). At the time of   
					your order, Tektronix will determine the part number revision level needed for   
					your product, based on the information you provide.   
					5-1   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Mechanical Parts List   
					Module Servicing   
					Modules can be serviced by selecting one of the following three options. Contact   
					your local Tektronix service center or representative for repair assistance.   
					Module Exchange. In some cases you may exchange your module for a reman-   
					ufactured module. These modules cost significantly less than new modules and   
					meet the same factory specifications. For more information about the module   
					exchange program, call 1-800-TEK-WIDE, extension 6630.   
					Module Repair and Return. You may ship your module to us for repair, after which   
					we will return it to you.   
					New Modules. You may purchase replacement modules in the same way as other   
					replacement parts.   
					Using the Replaceable Parts List   
					This section contains a list of the mechanical and/or electrical components that   
					are replaceable for the instrument. Use this list to identify and order replacement   
					parts. The following table describes each column in the parts list.   
					Parts List Column Descriptions   
					Column Column Name   
					Description   
					1 
					Figure & Index Number   
					Items in this section are referenced by figure and index numbers to the exploded view   
					illustrations that precede the list   
					2 
					Tektronix Part Number   
					Serial Number   
					Use this part number when ordering replacement parts from Tektronix   
					3 and 4   
					Column three indicates the serial number at which the part was first effective. Column four   
					indicates the serial number at which the part was discontinued. No entries indicates the part is   
					good for all serial numbers   
					5 
					6 
					Qty   
					This indicates the quantity of parts used   
					Name & Description   
					An item name is separated from the description by a colon (:). Because of space limitations, an   
					item name may sometimes appear as incomplete. Use the U.S. Federal Catalog handbook   
					H6-1 for further item name identification   
					7 
					8 
					Mfr. Code   
					This indicates the code of the actual manufacturer of the part   
					This indicates the actual manufacturer’s or vendor’s part number   
					Mfr. Part Number   
					Abbreviations   
					Abbreviations conform to American National Standard ANSI Y1.1--1972.   
					5-2   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Mechanical Parts List   
					Mfr. Code to Manufacturer   
					Cross Index   
					The table titled Manufacturers Cross Index shows codes, names, and addresses   
					of manufacturers or vendors of components listed in the parts list.   
					Manufacturers cross index   
					Mfr.   
					code   
					Manufacturer   
					Address   
					City, state, zip code   
					06383   
					06915   
					0B445   
					0KB01   
					0KB05   
					PANDUIT CORP   
					17303 RIDGELAND AVE   
					5825 N TRIPP AVE P.O. BOX 804238   
					312 EAST MAIN STREET   
					810 SE SHERMAN   
					TINLEY PARK, IL 60477-3048   
					CHICAGO, IL 60646   
					RICHCO   
					ELECTRI-CORD MFG CO INC   
					STAUFFER SUPPLY CO   
					NORTH STAR NAMEPLATE INC   
					WESTFIELD, PA 16950   
					PORTLAND, OR 97214-4657   
					HILLSBORO, OR 97124-6474   
					LIBERTY LAKE, WA 99019   
					DELAWARE WATER GAP, PA 18327-0650   
					5750 NE MOORE COURT   
					23201 E APPLEWAY   
					1WNW6 ACCRA-FAB INC   
					30817   
					LAIRD TECHNOLOGIES WORLD   
					ONE SHIELDING WAY PO BOX 650   
					HEADQUARTERS   
					NEC ELECTRONICS INC   
					CHOMERICS INC   
					4T165   
					57003   
					7X318   
					80009   
					2880 SCOTT BLVD PO BOX 58062   
					23839 S. BANNING BLVD   
					SANTA CLARA, CA 95052-2554   
					CARSON, CA 90745   
					KASO PLASTICS INC   
					TEKTRONIX INC   
					5720-C NE 121ST AVE, STE 110   
					VANCOUVER, WA 98682   
					BEAVERTON, OR 97077-0001   
					14150 SW KARL BRAUN DR   
					PO BOX 500   
					80640   
					ARTESYN TECHNOLOGY POWER   
					CONVERSION   
					7 ELKIN ST   
					SOUTH BOSTON, MA 02127-1615   
					83285   
					STEWARD INC   
					1200 E 36TH ST~PO BOX 510   
					10B VAN DYKE AVENUE   
					10156 TORINO   
					CHATTANOOGA, TN 37401-0510   
					NEW BRUNSWICK, NJ 08901   
					VAICENTALLO 62/456 ITALY   
					HILLSBORO, OR 97124   
					S3109   
					FELLER U.S. CORPORATION   
					PATELEC-CEM   
					TK1373   
					TK2548   
					TK6121   
					TK6159   
					TK6253   
					TK6433   
					XEROX CORPORATION   
					TUMBLER CORP   
					7431 EVERGREEN PARKWAY   
					4241 BUSINESS CENTER DR   
					2900 NW 229TH AVE   
					FREMONT, CA 94538   
					RAD TECHNOLOGIES   
					VOLEX INC   
					HILLSBORO, OR 97124   
					646 CARRIBEAN DR   
					SUNNYVALE, CA 94089-1108   
					I-SHENG ELECTRIC WIRE & PLASTIC   
					(KUNSHAN) CO LTD.   
					289 ZING YANG ZHONG RD   
					TECHNICAL DEVELOPMENT ZONE JIANGSU,   
					CHINA CN   
					TK6547   
					SOLECTRON MILPITAS INC   
					1177 GIBRALTAR DRIVE SOLECTRON   
					BUILDING 9   
					MILPITAS, CA 95035   
					TK6612   
					TK6619   
					HANACO MANUFACTURING COMPANY   
					ACE MOLD (SHANGHAI) COMPANY LTD   
					3965 SCHAEFER AVENUE   
					CHINO, CA 91710   
					318 NORTH ROUND-THE-CITY ROAD   
					SHANGHAI COMPREHENSIVE INDUSTRIAL   
					DEVELOPMENT ZONE, FENGXIAN SHANGHAI,   
					CN   
					TK6709   
					SOLECTRON (SHANGHAI)   
					798 YING LUN RO AD~WALGAOQIAO FTZ   
					PUDONG, SHANGHAI, CN CN 200131   
					5-3   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Mechanical Parts List   
					Exploded Views   
					Figures 5--1 through 5--5 on the following pages show the module-level   
					exploded views of the DPO4000 series oscilloscope. The adjacent page is the list   
					of components for that exploded view, indexed by the numbers in the figure.   
					3 
					2 
					4 
					4 
					1 
					5 
					Figure 5-1: Exploded view, case   
					5-4   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Mechanical Parts List   
					Replaceable Parts List   
					Fig. &   
					index   
					number   
					Tektronix part Serial no. Serial no.   
					discont’d   
					Mfr.   
					code   
					number   
					effective   
					Qty Name & description   
					Mfr. part number   
					5-1   
					-1   
					Case   
					200-4908-00   
					202-0441-00   
					211-1174-00   
					1 
					1 
					4 
					COVER, FRONT, PROTECTIVE   
					TK6710 200-4908-00   
					80009 202-0441-00   
					0KB01 211-1174-00   
					-2   
					CASE, REAR   
					-3   
					SCREW,MACHINE; W/HEAVY PATCH THREADLOCKING   
					MATERIAL; 6-32 X 0.312 L,PNH,STL CAD PLT,T15   
					-4   
					-5   
					335-1186-00   
					348-1798-00   
					1 
					2 
					MARKER,IDENT;LABEL,REAR PANEL   
					FOOT, FRONT   
					80009   
					80009   
					335-1186-00   
					348-1798-00   
					5-5   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Mechanical Parts List   
					6 
					5 
					4 
					7 
					8 
					3 
					9 
					2 
					1 
					10   
					15   
					Flex Circuit   
					J13   
					11   
					20   
					13   
					12   
					13   
					19   
					14   
					16   
					18   
					17   
					Figure 5-2: Exploded view: front panel and display, 4-channel instruments   
					5-6   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Mechanical Parts List   
					Replaceable Parts List   
					Fig. &   
					index   
					number   
					Tektronix part Serial no. Serial no.   
					discont’d   
					Mfr.   
					code   
					number   
					effective   
					Qty Name & description   
					Mfr. part number   
					5-2   
					-1   
					Front Panel   
					202-0432-00   
					335-1187-00   
					335-1191-00   
					335-1193-00   
					366-0860-01   
					366-0859-01   
					366-0861-01   
					366-0865-00   
					358-0883-00   
					366-0867-00   
					335-1467-00   
					335-1466-00   
					335-1465-00   
					335-1464-00   
					335-1463-00   
					335-1462-00   
					335-1461-00   
					335-1460-00   
					335-1459-00   
					335-1458-00   
					335-1183-00   
					200-4418-00   
					1 
					1 
					1 
					1 
					5 
					6 
					2 
					1 
					1 
					1 
					1 
					1 
					1 
					1 
					1 
					1 
					1 
					1 
					1 
					1 
					1 
					1 
					FRONT CASE ASSEMBLY   
					80009   
					80009   
					80009   
					80009   
					80009   
					80009   
					80009   
					80009   
					202-0432-00   
					335-1187-00   
					335-1191-00   
					335-1193-00   
					366-0860-01   
					366-0859-01   
					366-0861-01   
					366-0865-00   
					-2   
					LABEL. FRONT PANEL, MODEL DPO4054   
					LABEL. FRONT PANEL, MODEL DPO4104   
					LABEL. FRONT PANEL, MODEL DPO4034   
					KNOB, .685 DIA, SOFT TOUCH   
					-3   
					-4   
					-5   
					-6   
					-7   
					-8   
					-9   
					KNOB, .470 DIA, SOFT TOUCH   
					KNOB, .925 DIA, SOFT TOUCH   
					KNOB, JOG   
					BUSHING (JOG SHUTTLE KNOB)   
					TK2565 358-0883-00   
					KNOB, SHUTTLE   
					80009   
					80009   
					80009   
					80009   
					80009   
					80009   
					80009   
					80009   
					80009   
					80009   
					80009   
					80009   
					7X318   
					366-0867-00   
					335-1467-00   
					335-1466-00   
					335-1465-00   
					335-1464-00   
					335-1463-00   
					335-1462-00   
					335-1461-00   
					335-1460-00   
					335-1459-00   
					335-1458-00   
					335-1183-00   
					200-4418-00   
					LABEL. FRONT PANEL, OVERLAY, RUSSIAN   
					LABEL. FRONT PANEL, OVERLAY, KOREAN   
					LABEL. FRONT PANEL, OVERLAY, STD CHINESE   
					LABEL. FRONT PANEL, OVERLAY, SIM CHINESE   
					LABEL. FRONT PANEL, OVERLAY, PORTUGUESE   
					LABEL. FRONT PANEL, OVERLAY, JAPANESE   
					LABEL. FRONT PANEL, OVERLAY, SPANISH   
					LABEL. FRONT PANEL, OVERLAY, GERMAN   
					LABEL. FRONT PANEL, OVERLAY, ITALIAN   
					LABEL. FRONT PANEL, OVERLAY, FRENCH   
					LABEL. FRONT PANEL, 4CH   
					10   
					11   
					COVER; OPTION KEY DOOR,0.7270 X 1.2250 X 0.310,PC/   
					ABS,FR110,SILVER GRAY   
					12   
					13   
					260-2798-00   
					211-1174-00   
					1 
					9 
					SWITCH,KEYPAD;ELASTOMERIC FRONT PANEL   
					80009   
					260-2798-00   
					SCREW,MACHINE; W/HEAVY PATCH THREADLOCKING   
					MATERIAL; 6-32 X 0.312 L,PNH,STL CAD PLT,T15   
					0KB01 211-1174-00   
					14   
					15   
					16   
					672-6132-01   
					335-1194-00   
					348-1860-00   
					1 
					1 
					2 
					FRONT PANEL BOARD, 4 CH   
					80009   
					80009   
					57003   
					672-6132-00   
					335-1194-00   
					MARKER, FRONT PANEL, I/O   
					GASKET, ELEC; CONDUCTIVE URETHANE FOAM   
					77-03-3150-09600-   
					AP   
					17   
					18   
					19   
					20   
					441-2378-00   
					650-4715-00   
					174-5014-00   
					211-1174-00   
					1 
					1 
					1 
					4 
					CHASSIS;FRONT; SAFETY CONTROLLED   
					ASSEMBLY, LCD DISPLAY   
					TK6709 441-2378-00   
					80009   
					80009   
					650-4715-00   
					174-5014-00   
					CABLE ASSY,SP;DISPLAY DATA,FLAT FLEX   
					SCREW,MACHINE; W/HEAVY PATCH THREADLOCKING   
					MATERIAL; 6-32 X 0.312 L,PNH,STL CAD PLT,T15   
					0KB01 211-1174-00   
					5-7   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Mechanical Parts List   
					1 
					6 
					7 
					2 
					5 
					3 
					4 
					Figure 5-3: Exploded view: front panel 2-channel instruments   
					5-8   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Mechanical Parts List   
					Replaceable Parts List   
					Fig. &   
					index   
					number   
					Tektronix part Serial no. Serial no.   
					number   
					Mfr.   
					code   
					effective   
					discont’d   
					Qty Name & description   
					Mfr. part number   
					5-3   
					-1   
					-2   
					-3   
					-4   
					-5   
					-6   
					-7   
					Front Panel   
					335-1189-00   
					335-1185-00   
					260-2812-00   
					672-6133-01   
					200-4906-00   
					366-0860-01   
					366-0859-01   
					1 
					1 
					1 
					1 
					2 
					3 
					4 
					LABEL. FRONT PANEL, MODEL DPO4032   
					80009   
					80009   
					335-1189-00   
					335-1185-00   
					MARKER,IDENT;LABEL,FRONT PANEL,2CH   
					SWITCH,KEYPAD;ELASTOMERIC FRONT PANEL   
					CIRCUIT BOARD; FRONT PANEL;2 CHANNEL   
					COVER;BNC FILLER   
					TK6612 260-2812-00   
					80009 672-6133-01   
					TK6433 200-4906-00   
					KNOB, .685 DIA, SOFT TOUCH   
					80009   
					80009   
					366-0860-01   
					366-0859-01   
					KNOB, .470 DIA, SOFT TOUCH   
					5-9   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Mechanical Parts List   
					2 
					1 
					3 
					2 
					10   
					4 
					2 
					8 
					5 
					9 
					2 
					I/O board assemly   
					J310   
					7 
					Main board   
					J780   
					Main board   
					J850   
					6 
					Figure 5-4: Exploded view: power supply and I/O board   
					5-10   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Mechanical Parts List   
					Replaceable Parts List   
					Fig. &   
					index   
					number   
					Tektronix part Serial no. Serial no.   
					discont’d   
					Mfr.   
					code   
					number   
					effective   
					Qty Name & description   
					Mfr. part number   
					5-4   
					-1   
					Main Board   
					407-5019-00   
					211-0722-00   
					211-1213-00   
					119-7223-00   
					1 
					BRACKET;POWER SUPPLY COVER   
					TK6709 407-5019-00   
					-2   
					19   
					1 
					SCREW,MACHINE; 6-32 X 0.250,PNH,STL,CDPL,T-15 TORX 0KB01 211-0722-00   
					-3   
					SCREW, METRIC 5 X 10MM, FLATHEAD, STL, ZINC, POSI   
					0KB01 211-1213-00   
					TK6159 119-7223-00   
					-4   
					1 
					FAN,TUBEAXIAL; 12VDC,0.09A,1.4W,19   
					CFM,4000RPM,31DBA,60MM X 60MM X 25MM,W/7IN LEADS   
					& CONNECTOR,TACH OUT   
					-5   
					-6   
					-7   
					-8   
					-9   
					10   
					650-4939-00   
					366-0844-00   
					672-6131-01   
					335-1213-00   
					335-1214-00   
					337-4378-00   
					1 
					1 
					1 
					1 
					1 
					1 
					POWER SUPPLY SUB ASSEMBLY;DOMESTIC   
					PUSH BUTTON;POWER; SAFETY CONTROLLED   
					I/O BOARD   
					80009   
					650-4939-00   
					TK6619 366-0844-00   
					80009   
					80009   
					80009   
					80009   
					672-6131-01   
					335-1213-00   
					335-1214-00   
					337-4378-00   
					MARKER,IDENT;LABEL,REAR PANEL,LEFT   
					MARKER,IDENT;LABEL,REAR PANEL, RIGHT   
					SHIELD;INSULATOR,ELECTRICAL   
					5-11   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Mechanical Parts List   
					Display assembly   
					to Inverter board   
					assembly   
					1 
					7 
					CN2   
					2 
					Main board   
					Inverter board   
					assembly   
					3 
					4 
					CN1   
					J570   
					J571   
					5 
					6 
					9 
					Main board   
					J930   
					8 
					I/O board   
					assemly   
					J610   
					Main board   
					Figure 5-5: Exploded view: mainboard, inverter board, and side fans   
					5-12   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Mechanical Parts List   
					Replaceable Parts List   
					Fig. &   
					index   
					number   
					Tektronix part Serial no. Serial no.   
					number   
					Mfr.   
					code   
					effective   
					discont’d   
					Qty Name & description   
					Mfr. part number   
					5-5   
					-1   
					Main Board   
					650-4937-00   
					119-6806-00   
					1 
					1 
					FAN TRAY ASSEMBLY   
					80009   
					4T165   
					650-4937-00   
					104PW191   
					-2   
					POWER SUPPLY; INVERTER,1200VAC NO-LOAD & 600VAC   
					@ 85KOHM LOADED OUTPUT   
					-3   
					-4   
					-5   
					-6   
					211-1143-00   
					407-5113-00   
					174-5015-00   
					671-5843-00   
					671-5959-00   
					671-5960-00   
					671-5961-00   
					211-0722-00   
					2 
					1 
					1 
					1 
					1 
					1 
					1 
					6 
					SCREW,PT; K35-1.57,PAN HEAD,T-15   
					0KB01   
					211-1143-00   
					407-5113-00   
					174-5015-00   
					671-5843-00   
					671-5959-00   
					671-5960-00   
					671-5961-00   
					211-0722-00   
					BRACKET,PCB;BACKLIGHT BOARD   
					TK6433   
					80009   
					CABLE ASSY,SP;INVERTER POWER,FLAT FLEX   
					CIRCUIT BD ASSY; MAIN BOARD, (DPO4104)   
					CIRCUIT BD ASSY;350MHZ 4 CHANNEL, (DPO4054)   
					CIRCUIT BD ASSY;350MHZ 2 CHANNEL, (DPO4034)   
					CKT BD ASSY; 2 CHANNEL, (DPO4032)   
					TK6547   
					TK6547   
					TK6547   
					TK6547   
					-7   
					SCREW,MACHINE; 6-32 X 0.250,PNH,STL,CDPL,T-15 TORX 0KB01   
					DR   
					-8   
					-9   
					174-5014-00   
					211-1174-00   
					1 
					CABLE ASSY,SP;DISPLAY DATA,FLAT FLEX   
					80009   
					174-5014-00   
					211-1174-00   
					13   
					SCREW,MACHINE; W/HEAVY PATCH THREADLOCKING   
					MATERIAL; 6-32 X 0.312 L,PNH,STL CAD PLT,T15   
					5-13   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Mechanical Parts List   
					Replaceable Parts List   
					Fig. &   
					index   
					number   
					Tektronix   
					part number   
					Serial no. Serial no.   
					effective discont’d   
					Mfr.   
					code   
					Qty Name & description   
					STANDARD ACCESSORIES   
					A c c e s s o r ie s   
					Mfr. part number   
					- - -   
					161-0104-05   
					161-0104-06   
					161-0104-07   
					161-0167-00   
					161-0306-00   
					161-0066-09   
					161-0298-00   
					161-0066-00   
					- - - - - - - -   
					1 
					1 
					1 
					1 
					1 
					1 
					1 
					1 
					1 
					CA ASSY,PWR:3,1.0MM SQ,250V/10A,2.5 M   
					(OPTION A3 - AUSTRALIAN)   
					TK1373 161-0104-05   
					CA ASSY,PWR:3,1.0MM SQ,250V/10A,2.5 M   
					(OPTION A1 - EUROPEAN)   
					S3109   
					1923   
					CA ASSY,PWR:3,1.0MM SQ,240V/10A,2.5 M   
					(OPTION A2 - UNITED KINGDOM)   
					TK6121 21072WH518.2   
					CA ASSY,PWR:3,0.75MM SQ,250V/10A,2.5 M   
					(OPTION A5 - SWITZERLAND)   
					0B445   
					161-0167-00   
					CA ASSY,PWR:3,1.0MM SQ,250V/10A,2.5 M   
					(OPTION A10 - CHINA)   
					TK6253 92-2637-250BKH   
					TK6121 2171H518.2   
					CA ASSY,PWR:3,1.0MM SQ,250V/10A,2.5 M   
					(OPTION A11 - INDIA)   
					CA ASSY,PWR:3,1.0MM SQ,250V/10A,2.5 M   
					(OPTION A6 - JAPAN)   
					0B445   
					S3109   
					161-0298-00   
					161-0066-00   
					CABLE ASSY,PWER,:3,18 AWG,92 L   
					(STANDARD CABLE)   
					POUCH, PLASTIC:POUCH   
					TK2582 TK1441 BLACK   
					CORDURA   
					5-14   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Index   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Download from Www.Somanuals.com. All Manuals Search And Download.   
				Index   
					A 
					I 
					acquisition system, description, 2-2   
					adjustment procedure, 3-2   
					factory, 3-3   
					I/O board removal, 4-10   
					inspection   
					exterior, 4-3   
					adjustments, 3-1   
					application modules, ix   
					interior, 4-4   
					inspection and cleaning   
					exterior, 4-2   
					flat panel display, 4-2   
					interior, 4-2   
					instrument calibration, 3-1   
					internal modules, 4-8   
					inverter board removal, 4-10   
					I/O board connections, 2-2   
					C 
					cabinet, 4-7   
					calibration due date, 3-1   
					calibration switch, 3-3   
					circuit description, 2-1–2-4   
					cleaning, 4-5   
					cleaning and inspection   
					exterior, 4-2, 4-3   
					flat panel display, 4-2   
					interior, 4-2   
					K 
					keypad, 4-8   
					connectivity, ix   
					L 
					lubrication, 4-3   
					D 
					description, circuit theory, 2-1–2-4   
					diagram, 2-1   
					display module removal, 4-11   
					display system, description, 2-2   
					documentation, ix   
					M 
					main board functions, 2-2   
					main board removal, 4-10   
					manual conventions, ix   
					manuals, ix   
					module interconnections, 2-1   
					E 
					electrostatic discharge, 4-1   
					precautions, 4-1–4-2   
					environmental characteristics, 4-16   
					P 
					packing, 4-16   
					power converter, description, 2-3   
					power supply, 2-2   
					power supply removal, 4-9   
					preventive maintenance, 4-1   
					probe compensation, 2-3   
					procedures   
					F 
					factory adjustment, 3-2   
					fans removal, 4-10   
					flat panel display, cleaning, 4-2   
					flowcharts, troubleshooting, 4-13   
					front case removal, 4-10   
					front panel removal, 4-11   
					front panel, description, 2-3   
					flat panel cleaning, 4-2   
					inspection and cleaning, 4-1–4-5   
					interior cleaning, 4-5   
					processor system, description, 2-3   
					Index-1   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Index   
					replaceable part, definition, ix   
					R 
					rear case removal, 4-9   
					removal procedures, 4-6, 4-8   
					display module, 4-11   
					fans, 4-10   
					T 
					trigger system, description, 2-2   
					trim, 4-7   
					troubleshoot, 4-12   
					front case, 4-10   
					front panel, 4-11   
					I/O board, 4-10   
					inverter board, 4-10   
					main board, 4-10   
					power supply, 4-9   
					rear case, 4-9   
					U 
					user manual, 1-1   
					Index-2   
					DPO4000 Series Service Manual   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 |