Tektronix Photo Scanner 071 1844 00 User Manual

Service Manual  
DPO4000 Series  
Digital Phosphor Oscilloscopes  
071-1844-00  
This document applies to firmware version 1.00  
and above.  
Warning  
The servicing instructions are for use by qualified  
personnel only. To avoid personal injury, do not  
perform any servicing unless you are qualified to  
do so. Refer to all safety summaries prior to  
performing service.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Warranty 16  
Tektronix warrants that the product will be free from defects in materials and workmanship for a period of three  
(3) years from the date of original purchase from an authorized Tektronix distributor. If the product proves  
defective during this warranty period, Tektronix, at its option, either will repair the defective product without  
charge for parts and labor, or will provide a replacement in exchange for the defective product. Batteries are  
excluded from this warranty. Parts, modules and replacement products used by Tektronix for warranty work may  
be new or reconditioned to like new performance. All replaced parts, modules and products become the property  
of Tektronix.  
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration  
of the warranty period and make suitable arrangements for the performance of service. Customer shall be  
responsible for packaging and shipping the defective product to the service center designated by Tektronix,  
shipping charges prepaid, and with a copy of customer proof of purchase. Tektronix shall pay for the return of the  
product to Customer if the shipment is to a location within the country in which the Tektronix service center is  
located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for  
products returned to any other locations.  
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate  
maintenance and care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage  
resulting from attempts by personnel other than Tektronix representatives to install, repair or service the product;  
b) to repair damage resulting from improper use or connection to incompatible equipment; c) to repair any  
damage or malfunction caused by the use of non-Tektronix supplies; or d) to service a product that has been  
modified or integrated with other products when the effect of such modification or integration increases the time  
or difficulty of servicing the product.  
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THE PRODUCT IN LIEU OF ANY  
OTHER WARRANTIES, EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY  
IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.  
TEKTRONIX’ RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND  
EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY.  
TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL,  
OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS  
ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table of Contents  
General Safety Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Service Safety Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Environmental Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
iii  
v
vii  
Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Manual Conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Related Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
ix  
ix  
ix  
Operating Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
1--1  
Theory of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Block Diagrams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
I/O Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Main Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Front-Panel Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
2--1  
2-1  
2-2  
2-2  
2-2  
2-3  
Adjustment Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Required Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Overview of the Adjustment Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Factory Adjustment Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
3--1  
3-1  
3-2  
3-3  
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Preventing ESD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Inspection and Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
General Care . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Interior Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Exterior Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Overview of Removal Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Removal Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Troubleshooting Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Unpacking and Repacking Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
4--1  
4-1  
4-1  
4-2  
4-2  
4-2  
4-6  
4-8  
4-12  
4-13  
4-16  
Mechanical Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Parts Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Using the Replaceable Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
5--1  
5-1  
5-2  
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Index--1  
i
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table of Contents  
List of Tables  
Table 4--1: External inspection check list . . . . . . . . . . . . . . . . . . . . . .  
Table 4--2: Internal inspection check list . . . . . . . . . . . . . . . . . . . . . .  
4--3  
4--4  
List of Figures  
Figure 2--1: DPO4000 series block diagram . . . . . . . . . . . . . . . . . . . .  
Figure 4--1: Locator for trim and cabinet removal . . . . . . . . . . . . . . .  
Figure 4--2: Locator for internal modules . . . . . . . . . . . . . . . . . . . . . .  
Figure 4--3: Primary troubleshooting procedure . . . . . . . . . . . . . . . .  
Figure 4--4: AC power supply troubleshooting procedure . . . . . . . . .  
Figure 4--5: Module isolation troubleshooting procedure . . . . . . . . .  
Figure 5--1: Exploded view, case . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
2--1  
4--7  
4--8  
4--13  
4--14  
4--15  
5--4  
Figure 5--2: Exploded view: front panel and display,  
4-channel instruments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
5--6  
5--8  
Figure 5--3: Exploded view: front panel 2-channel instruments . . . .  
Figure 5--4: Exploded view: power supply and I/O board . . . . . . . .  
5--10  
Figure 5--5: Exploded view: mainboard, inverter board,  
and side fans . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
5--12  
ii  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
General Safety Summary  
Review the following safety precautions to avoid injury and prevent damage to  
this product or any products connected to it.  
To avoid potential hazards, use this product only as specified.  
Only qualified personnel should perform service procedures.  
Use Proper Power Cord. Use only the power cord specified for this product and  
certified for the country of use.  
To Avoid Fire or  
Personal Injury  
Connect and Disconnect Properly. Do not connect or disconnect probes or test  
leads while they are connected to a voltage source.  
Ground the Product. This product is grounded through the grounding conductor  
of the power cord. To avoid electric shock, the grounding conductor must be  
connected to earth ground. Before making connections to the input or output  
terminals of the product, ensure that the product is properly grounded.  
Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings  
and markings on the product. Consult the product manual for further ratings  
information before making connections to the product.  
The inputs are not rated for connection to mains or Category II, III, or IV  
circuits.  
Connect the probe reference lead to earth ground only.  
Do not apply a potential to any terminal, including the common terminal, that  
exceeds the maximum rating of that terminal.  
Do Not Operate Without Covers. Do not operate this product with covers or panels  
removed.  
Do Not Operate With Suspected Failures. If you suspect there is damage to this  
product, have it inspected by qualified service personnel.  
Avoid Exposed Circuitry. Do not touch exposed connections and components  
when power is present.  
Do Not Operate in Wet/Damp Conditions.  
Do Not Operate in an Explosive Atmosphere.  
Keep Product Surfaces Clean and Dry.  
Provide Proper Ventilation. Refer to the manual’s installation instructions for  
details on installing the product so it has proper ventilation.  
iii  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
General Safety Summary  
Terms in this Manual  
These terms may appear in this manual:  
WARNING. Warning statements identify conditions or practices that could result  
in injury or loss of life.  
CAUTION. Caution statements identify conditions or practices that could result in  
damage to this product or other property.  
Symbols and Terms  
on the Product  
These terms may appear on the product:  
H
H
H
DANGER indicates an injury hazard immediately accessible as you read the  
marking.  
WARNING indicates an injury hazard not immediately accessible as you  
read the marking.  
CAUTION indicates a hazard to property including the product.  
The following symbols may appear on the product:  
Protective Ground  
(Earth) Terminal  
CAUTION  
Refer to Manual  
Mains Connected  
ON (Power)  
Push button OUT  
Chassis Ground  
Push button  
IN  
iv  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Service Safety Summary  
Only qualified personnel should perform service procedures. Read this Service  
Safety Summary and the General Safety Summary before performing any service  
procedures.  
Do Not Service Alone. Do not perform internal service or adjustments of this  
product unless another person capable of rendering first aid and resuscitation is  
present.  
Disconnect Power. To avoid electric shock, switch off the instrument power, then  
disconnect the power cord from the mains power.  
Use Care When Servicing With Power On. Dangerous voltages or currents may  
exist in this product. Disconnect power, remove battery (if applicable), and  
disconnect test leads before removing protective panels, soldering, or replacing  
components.  
To avoid electric shock, do not touch exposed connections.  
v
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Service Safety Summary  
vi  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Environmental Considerations  
This section provides information about the environmental impact of the  
product.  
Observe the following guidelines when recycling an instrument or component:  
Product End-of-Life  
Handling  
Equipment Recycling. Production of this equipment required the extraction and  
use of natural resources. The equipment may contain substances that could be  
harmful to the environment or human health if improperly handled at the  
product’s end of life. In order to avoid release of such substances into the  
environment and to reduce the use of natural resources, we encourage you to  
recycle this product in an appropriate system that will ensure that most of the  
materials are reused or recycled appropriately.  
The symbol shown to the left indicates that this product  
complies with the European Union’s requirements  
according to Directive 2002/96/EC on waste electrical and  
electronic equipment (WEEE). For information about  
recycling options, check the Support/Service section of the  
Tektronix Web site (www.tektronix.com).  
Mercury Notification. This product uses an LCD backlight lamp that contains  
mercury. Disposal may be regulated due to environmental considerations. Please  
contact your local authorities or, within the United States, the Electronics  
This product has been classified as Monitoring and Control equipment, and is  
outside the scope of the 2002/95/EC RoHS Directive. This product is known to  
contain lead, cadmium, mercury, and hexavalent chromium.  
Restriction of Hazardous  
Substances  
vii  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Environmental Considerations  
viii  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Preface  
This Service manual provides information to troubleshoot, disassemble, and  
replace parts on the DPO4000 Series Digital Phosphor Oscilloscopes.  
Manual Conventions  
This manual uses certain conventions that you should become familiar with  
before attempting service.  
Modules  
Throughout this manual, any replaceable component, assembly, or part is  
referred to by the term module. A module is composed of electrical and  
mechanical assemblies, circuit cards, interconnecting cables, and user-accessible  
controls.  
Replaceable Parts  
This manual refers to any field-replaceable assembly or mechanical part  
specifically by its name or generically as a replaceable part. In general, a  
replaceable part is any circuit board or assembly, such as the hard disk drive, or a  
mechanical part, such as the I/O port connectors, that is listed in the replaceable  
parts list.  
Safety  
Symbols and terms related to safety appear in the Service Safety Summary found  
at the beginning of this manual.  
Related Documentation  
To read about  
Use these documents  
Installation and Operation  
DPO4000 Series Digital Phosphor Oscilloscopes User  
Manual (available in 11 languages)  
Specifications and Performance DPO4000 Technical Reference (PDF only)  
Programmer Commands DPO4000 Programmer Manual (PDF only)  
Analysis and Connectivity Tools Getting Started with OpenChoicet Solutions Manual  
Installing and testing applica-  
tions modules  
DPO4000 Series Application Modules Installation Instructions  
Manual  
ix  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Preface  
x
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operating Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operating Information  
For information on installing and operating your DPO4000 Series Digital  
Phosphor Oscilloscope, please refer to the DPO4000 Series Digital Phosphor  
Oscilloscopes User Manual. This manual is available on the Web at  
1-1  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operating Information  
1-2  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Theory of Operation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Theory of Operation  
This chapter describes the electrical operation of the oscilloscope to the module  
level. Figure 2--1 shows the oscilloscope module interconnections.  
Block Diagrams  
Display  
back light  
inverter  
power  
LCD  
Display  
Aux Out  
J170  
J930  
J680  
AC Line  
Power  
supply  
Front panel  
J1  
J100  
I/O Board  
J440  
Input ports for  
application  
modules  
J0480  
J610  
Compact  
Flash Port  
J420, J421  
Main board  
J120  
J130  
J570  
J571  
J850  
USB  
J510  
Bezel  
buttons  
3 Fans  
Attenuator circuit  
Ethernet  
connector  
Ch 1  
Ch 1  
Ch 2  
Ch 2  
Ch 3  
Ch 4  
EXT  
EXT 4-channel oscilloscopes  
2-channel oscilloscopes  
Figure 2-1: DPO4000 series block diagram  
2-1  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Theory of Operation  
Power Supply  
The power supply converts AC line voltage to +12V to power all internal  
circuits.  
I/O Board  
The I/O board contains USB ports, Compact Flash port, an Ethernet connector,  
TekLink, XGA Video, and a Trigger Out port.  
Main Board  
The main board module contains the following functions:  
Acquisition System  
The acquisition system begins with the analog signal path and ends with a  
digitized signal in memory. The signal enters a channel input, and then passes  
through an attenuator and preamplifier. The analog signal from each preamplifier  
goes through a digitizer, a time base controller, and then into acquisition  
memory. The analog signal from each preamplifier is also distributed to a trigger  
circuit.  
Trigger System  
Display System  
The trigger system processes the analog signals from the preamplifiers and sends  
trigger information to the time base controller. Advanced trigger functions are  
enabled only when the appropriate application modules and supporting software  
are installed.  
The display system combines live waveform data from acquisition memory with  
menus and text and stores this information in display memory. It then uses this  
data to refresh the XGA display module (LCD).  
2-2  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Theory of Operation  
Processor System  
The processor system contains a 440EP Power PC microprocessor that controls  
the entire instrument. The processor system also contains FLASH ROM, system  
RAM, and interfaces to USB ports and a Compact flash slot.  
Power Converter  
The power converter receives +12V from the main power supply and generates  
voltages for the analog and digital circuitry on the main board.  
Front-Panel Board  
The front-panel board contains a microprocessor that reads the front-panel  
buttons and controls, and then sends this information to the processor system on  
the main board. The front-panel board also generates the probe compensation  
output signal and provides an interface to the application modules.  
2-3  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Theory of Operation  
2-4  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Adjustment Procedures  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Adjustment Procedures  
This chapter contains the factory adjustment procedures for the DPO4000 series  
oscilloscopes. Only qualified personnel should perform adjustment procedures.  
Read the Service Safety Summary and the General Safety Summary before  
performing any service procedures.  
NOTE. The voltage references inside the oscilloscope are very stable over time  
and should not require routine adjustment. Before performing any procedure in  
this chapter, do the Performance Verification procedures to check whether the  
oscilloscope meets specifications. See the DPO4000 Technical Reference.  
Successful completion of this adjustment procedure automatically updates the  
instrument Calibration Due date and time. (Completion of the Performance  
Verification procedure does not update the instrument Calibration Due date and  
time.)  
Required Equipment  
The following equipment, or a suitable equivalent, is required to complete these  
procedures.  
Description  
Minimum requirements  
Example  
DC voltage source  
50 mV to 70 V, 0.1% amplitude Wavetek 9500 Oscilloscope  
accuracy  
Calibrator with one 9510  
Output Module  
Sine Generator  
Edge Generator  
50 Ω BNC cable  
1 GHz  
Wavetek 9500 with 9530  
output modules  
1 kHz with <50 ps ch-ch skew  
BNC male to BNC male, 10 in Tektronix part number  
(25 cm) long 012-0208-00  
Adjustment tool  
Non conducting shaft with diame- Tektronix part number  
ter 2.5 mm (0.1 in) 003-1433-00  
3-1  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Adjustment Procedures  
Overview of the Adjustment Process  
Before performing adjustment procedures, you must warm up the oscilloscope at  
least ten minutes in an ambient temperature between 20 °C and 30 °C. Adjust-  
ments performed before warm-up or outside this temperature range may result in  
poor performance.  
The factory adjustment procedure consists of a series of steps; as you move  
through these steps, the oscilloscope display provides instructions that describe  
the specific input signal requirements for each step. If the the oscilloscope passes  
the step, it moves on to the next step. If the oscilloscope fails, you can repeat the  
step or choose to abort the procedure.  
NOTE. Do not turn any knobs or push any front-panel buttons other than the  
Next Step or Previous Step buttons during the adjustment procedure. Doing so  
will cause the oscilloscope to abort the adjustment procedure. The oscilloscope  
uses the previous calibration constants if the adjustment procedure is aborted.  
The screen does not display the actual oscilloscope settings (such as channel  
input impedance or vertical and horizontal settings) during the adjustment  
procedure. The oscilloscope automatically sets the instrument settings, but these  
settings may not read out correctly on the display.  
During some steps, the instrument may appear to be idle for several minutes  
while it is processing information internally.  
If the oscilloscope completes all steps in the procedure successfully, a “Pass”  
message is displayed and the new calibration constants take effect. If the  
oscilloscope does not pass the procedure, the oscilloscope reverts to the previous  
calibration settings.  
3-2  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Adjustment Procedures  
Factory Adjustment Procedure  
To perform the factory adjustment procedure, do these steps:  
1. Connect the oscilloscope to an AC power source.  
NOTE. You must connect the oscilloscope and the test equipment to the same AC  
power circuit. Connect the oscilloscope and test instruments to a common power  
strip if you are unsure of the AC power circuit distribution. Connecting the  
oscilloscope and test instruments to separate AC power circuits can result in  
offset voltages between the equipment, which can invalidate the adjustment  
procedure.  
2. Locate the hole in the rear panel that provides access to the CAL switch on  
the rear panel.  
CAL  
3. Insert the adjustment tool into the CAL switch hole. Use the tool to push and  
hold the CAL switch in while you power on the oscilloscope. Release the  
switch when the oscilloscope shows the power-on message with the  
oscilloscope graticule in the background.  
4. Push the Default Setup button.  
NOTE. Do not skip the recall Default Setup step. You must recall the factory  
settings in order to correctly start the factory adjustment procedure.  
5. Push the MENU OFF button.  
6. Push the UTILITY button.  
7. Push the System Config lower bezel button repeatedly to select Calibra-  
tion.  
3-3  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Adjustment Procedures  
8. Push the Factory lower bezel button.  
9. Push the OK Do Factory Calibration side bezel button to start the  
adjustment process.  
10. Connect signals from the DC voltage source as instructed by the oscilloscope  
display. After you connect the correct signal, push the OK Do Next Step  
side bezel button to continue the process.  
NOTE. The oscilloscope adjusts itself automatically using the reference signal  
from the DC voltage source. You do not need to make any adjustments.  
If you make an error, such as connecting the wrong input signal, you can repeat  
the last step by pushing the Return to Previous Screen side bezel button.  
11. Continue with the adjustment process until it is complete. You can abort the  
process at any time by pushing the MENU OFF button.  
12. Do the performance verification tests to verify that the adjustment procedure  
has correctly calibrated the oscilloscope. (See the DPO4000 Technical  
Reference.) Incorrect use of calibration equipment can cause the oscilloscope  
to pass the adjustment procedure but fail performance verification.  
3-4  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
This section contains the information needed to do periodic and corrective  
maintenance on the oscilloscope as well as repackaging instructions if you need  
to return the oscilloscope to Tektronix for service.  
Preventing ESD  
Before servicing this product, read the Safety Summary and Introduction at the  
front of the manual and the electrostatic discharge (ESD) information below.  
CAUTION. Static discharge can damage any semiconductor component in this  
oscilloscope.  
When performing any service that requires internal access to the oscilloscope,  
adhere to the following precautions to avoid damaging internal modules and their  
components due to electrostatic discharge.  
1. Minimize handling of static-sensitive circuit boards and components.  
2. Transport and store static-sensitive modules in their static protected  
containers or on a metal rail. Label any package that contains static-sensitive  
boards.  
3. Discharge the static voltage from your body by wearing a grounded antistatic  
wrist strap while handling these modules. Do service of static-sensitive  
modules only at a static-free work station.  
4. Do not place anything capable of generating or holding a static charge on the  
work station surface.  
5. Handle circuit boards by the edges when possible.  
6. Do not slide the circuit boards over any surface.  
7. Avoid handling circuit boards in areas that have a floor or work-surface  
covering capable of generating a static charge.  
Inspection and Cleaning  
Inspection and cleaning are done as preventive maintenance. Preventive  
maintenance, when done regularly, may prevent oscilloscope malfunction and  
enhance its reliability.  
4-1  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
Preventive maintenance consists of visually inspecting and cleaning the  
oscilloscope and using general care when operating it.  
How often you do maintenance depends on the severity of the environment in  
which the oscilloscope is used. A proper time to perform preventive maintenance  
is just before oscilloscope adjustment.  
General Care  
The cabinet helps keep dust out of the oscilloscope and should normally be in  
place when operating the oscilloscope.  
Flat Panel Display  
Cleaning  
The display is a soft plastic display and must be treated with care during  
cleaning.  
CAUTION. Improper cleaning agents or methods can damage the flat panel  
display.  
Avoid using abrasive cleaners or commercial glass cleaners to clean the display  
surface.  
Avoid spraying liquids directly on the display surface.  
Avoid scrubbing the display with excessive force.  
Clean the flat panel display surface by gently rubbing the display with a  
clean-room wipe (such as Wypall Medium Duty Wipes, #05701, available from  
Kimberly-Clark Corporation).  
Interior Cleaning  
Use a dry, low-velocity stream of air to clean the interior of the chassis. Use a  
soft-bristle, non-static-producing brush for cleaning around components. If you  
must use a liquid for minor interior cleaning, use a 75% isopropyl alcohol  
solution and rinse with deionized water.  
WARNING. To avoid injury, power down the instrument and disconnect it from  
line voltage before performing any procedure that follows.  
Exterior Cleaning  
Clean the exterior surfaces of the chassis with a dry lint-free cloth or a soft-  
bristle brush. If any dirt remains, use a cloth or swab dipped in a 75% isopropyl  
alcohol solution. Use a swab to clean narrow spaces around controls and  
connectors. Do not use abrasive compounds on any part of the chassis that may  
damage the chassis.  
Clean the On/Standby switch using a dampened cleaning towel. Do not spray or  
wet the switch directly.  
4-2  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
CAUTION. Avoid the use of chemical cleaning agents, which might damage the  
plastics used in this oscilloscope. Use only deionized water when cleaning the  
menu buttons or front-panel buttons. Use a 75% isopropyl alcohol solution as a  
cleaner and rinse with deionized water. Before using any other type of cleaner,  
consult your Tektronix Service Center or representative.  
Lubrication. There is no periodic lubrication required for this oscilloscope.  
Exterior Inspection  
Inspect the outside of the oscilloscope for damage, wear, and missing parts,  
using Table 4--1 as a guide. Immediately repair defects that could cause personal  
injury or lead to further damage to the oscilloscope.  
Table 4-1: External inspection check list  
Item  
Inspect for  
Repair action  
Cabinet, front panel,  
and cover  
Cracks, scratches, deformations, Repair or replace defective  
damaged hardware  
module  
Front-panel knobs  
Missing, damaged, or loose  
knobs  
Repair or replace missing or  
defective knobs  
Connectors  
Broken shells, cracked insulation, Repair or replace defective  
and deformed contacts. Dirt in  
connectors  
modules. Clear or wash out dirt.  
Carrying handle, and  
cabinet feet  
Correct operation  
Repair or replace defective  
module  
Accessories  
Missing items or parts of items,  
bent pins, broken or frayed  
Repair or replace damaged or  
missing items, frayed cables, and  
cables, and damaged connectors defective modules  
If the display is very dirty, moisten the wipe with distilled water or a 75%  
isopropyl alcohol solution and gently rub the display surface. Avoid using excess  
force or you may damage the plastic display surface.  
CAUTION. To prevent getting moisture inside the oscilloscope during external  
cleaning, use only enough liquid to dampen the cloth or applicator.  
4-3  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
Interior Inspection  
To access the inside of the oscilloscope for inspection and cleaning, refer to the  
Removal Procedures in this section.  
Inspect the internal portions of the oscilloscope for damage and wear, using  
Table 4--2 as a guide. Repair any defects immediately.  
If any circuit board is repaired or replaced, check Table 4--2 to see if it is  
necessary to adjust the oscilloscope.  
CAUTION. To prevent damage from electrical arcing, ensure that circuit boards  
and components are dry before applying power to the oscilloscope.  
Table 4-2: Internal inspection check list  
Item  
Inspect for  
Repair action  
Circuit boards  
Loose, broken, or corroded  
solder connections. Burned  
circuit boards. Burned, broken, or  
cracked circuit-run plating.  
Remove and replace damaged  
circuit board.  
Resistors  
Burned, cracked, broken, blis-  
tered condition.  
Remove and replace damaged  
circuit board.  
Solder connections  
Capacitors  
Cold solder or rosin joints.  
Resolder joint and clean with  
isopropyl alcohol.  
Damaged or leaking cases.  
Corroded solder on leads or  
terminals.  
Remove and replace damaged  
circuit board.  
Wiring and cables  
Chassis  
Loose plugs or connectors.  
Burned, broken, or frayed wiring. replace modules with defective  
wires or cables.  
Firmly seat connectors. Repair or  
Dents, deformations, and dam-  
aged hardware.  
Straighten, repair, or replace  
defective hardware.  
4-4  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
Cleaning Procedure — Interior. To clean the oscilloscope interior, do the following  
steps:  
1. Blow off dust with dry, low-pressure, deionized air (approximately 9 psi).  
2. Remove any remaining dust with a lint-free cloth dampened in isopropyl  
alcohol (75% solution) and rinsed with warm deionized water. (A cotton-  
tipped applicator is useful for cleaning in narrow spaces and on circuit  
boards.)  
STOP. If, after doing steps 1 and 2, a module is clean upon inspection, skip the  
following steps.  
If steps 1 and 2 do not remove all the dust or dirt, the oscilloscope may be spray  
washed using a solution of 75% isopropyl alcohol by doing the following steps:  
1. Gain access to the parts to be cleaned by removing easily accessible shields  
and panels (see Removal Procedures).  
2. Spray wash dirty parts with the isopropyl alcohol and wait 60 seconds for the  
majority of the alcohol to evaporate.  
3. Use hot (120 _F to 140 _F) deionized water to thoroughly rinse them.  
4. Dry all parts with low-pressure, deionized air.  
5. Dry all components and assemblies in an oven or drying compartment using  
low-temperature (125 _F to 150 _F) circulating air.  
4-5  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
Overview of Removal Procedures  
Refer to Figures 4--1 and 4--2 for the location of modules that you are removing  
or installing. If you are disassembling the instrument for cleaning, refer to the  
Inspection and Cleaning procedure on page 4--1 for instructions.  
List of Modules  
The Mechanical Parts List chapter provides a list of all replaceable modules.  
Any replaceable component, assembly, or part is referred to by the term module.  
WARNING. Only qualified personnel should perform service procedures. Before  
doing this or any other procedure in this manual, read the General Safety  
Summary and Service Safety Summary located at the beginning of this manual.  
Also, to prevent possible injury to service personnel or damage to electrical  
components, read Preventing ESD on page 4--1.  
Tools Required  
You will need the following tools to remove and replace all the modules in the  
instrument.  
H
Torque-limiting screwdriver, 12 in lb (1.3 N m) range with TORX T-15  
tips. (If you are using a magnetic screwdriver with interchangeable bits, the  
T-15 bit must be a long bit, at least 5 cm (2 in) overall length.)  
H
H
H
H
Screwdriver with 1  
8 inch flat blade  
@
Phillips screwdrivers with P0 and P2 tips  
3
1@ 6 inch nut driver  
Needle-nose pliers  
Use the following tables to remove the trim, cabinet, and internal modules.  
Trim, Cabinet, and Module  
Removal  
WARNING. Before doing any procedure in this subsection, disconnect the power  
cord from the line voltage source. Failure to do so could cause serious injury or  
death.  
4-6  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
Rear case  
Front foot  
Front cover  
Figure 4-1: Locator for trim and cabinet removal  
4-7  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
Power supply  
assembly  
Inverter board  
assembly  
Side fans  
assembly  
Display  
assembly  
I/O board  
assembly  
Main board  
Front Case  
assembly  
Front Panel  
assembly  
Figure 4-2: Locator for internal modules  
Removal Procedures  
These procedures assume you have access to the module that you are removing.  
Refer to the exploded drawings on pages 5--4 through 5--12 and to Figures 4--1  
and 4--2, on pages 4--7 and 4--8 to determine which trim and or modules to  
remove.  
CAUTION. When removing or installing the keypad, make sure you do not touch  
the switch contacts with your fingers. The oils in your fingers will degrade or  
damage the switch contacts. To help prevent damage to the keypad use cotton  
gloves when removing or installing the keyboard pad.  
4-8  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
CAUTION. To avoid damage to the front panel Standby/On switch assembly, do  
not set the Display module assembly on a work surface. Sliding the oscilloscope  
over the edge of the work surface could break off the On/Standby switch  
assembly.  
Rear Case  
Follow these steps to remove the rear case. Use a TORX T-15 screwdriver tip.  
See Figure 5--1 on page 5--4.  
1. Remove the 4 screws from the back and sides of the rear case.  
2. Grasp the case and pull outwards to remove it.  
Power Supply Module  
Follow these steps to remove the power supply module. You need to have  
previously removed the rear case. Use a TORX T-15 screwdriver tip. See  
Figure 5--4 on page 5--10.  
1. Remove the 7 screws that secure the power supply module to the chassis.  
2. Disconnect the power supply fan and connectors from the I/O board and the  
main board.  
3. Lift the power supply module off the chassis.  
Power Supply Fan. Follow these steps to remove the power supply fan assembly.  
You need to have previously removed the rear case and power supply assembly.  
Use a TORX T-15 and a P2 screwdriver tip. See Figure 5--4 on page 5--10.  
1. Remove the 5 screws that attach the power supply cover.  
2. Lift the power supply cover off.  
3. Use the P2 screwdriver tip and remove the 2 screws that attach the fan to the  
cover.  
Power Supply. Follow these steps to remove the power supply. You need to have  
previously removed the rear case and power supply assembly. Use a TORX T-15  
screwdriver tip. See Figure 5--4 on page 5--10.  
1. Remove the 5 screws that attach the power supply cover.  
2. Lift the power supply cover off.  
3. Remove the 4 screws that attach the power supply to the power supply  
bracket.  
4-9  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
I/O Board  
Follow these steps to remove the I/O board. You need to have previously  
removed the rear case and power supply module. Use TORX T-15 screwdriver  
tip. See Figure 5--4 on page 5--10.  
1. Remove the 5 screws that hold the I/O board to the chassis.  
2. Lift straight up to remove.  
Main Fans  
Follow these steps to remove the side fan assembly. You need to have previously  
removed the rear case, power supply assembly, and I/O board. Use a TORX T-15  
screwdriver tip. See Figure 5--5 on page 5--12.  
1. Remove the 6 screws that hold the side fans to the chassis.  
2. Disconnect the two fan connectors from the main board.  
3. Slide the fan assembly out and away from the chassis.  
Main Board  
Follow these steps to remove the main board. You need to have previously  
removed the rear case, power supply, I/O board, and side fans. Use a TORX T-15  
screwdriver tip. See Figure 5--5 on page 5--12.  
1. Disconnect the display cable from the main board.  
2. Disconnect the backlight cable from the inverter board.  
3. Remove 13 screws from the main board that connect it to the chassis.  
4. Lift the main board up to remove.  
Inverter board. Follow these steps to remove the inverter board from the main  
board. You must have previously removed the main board. Use a TORX T-15  
screwdriver tip. See Figure 5--5 on page 5--12.  
1. Disconnect the inverter board cable from the main board.  
2. Remove the inverter board from the plastic bracket.  
3. Remove the 2 screws that hold the inverter board bracket to the main board.  
4. Remove the inverter board bracket from the main board.  
4-10  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
Front Case Assembly  
Follow these steps to remove the front case assembly. You must have previously  
removed the Rear Case Assembly. See Figure 5--2 on page 5--6.  
1. Pull off 14 knobs (2 large, 5 medium, 6 small, and 1 rotary).  
2. Remove the front case by pulling back the board snaps and lifting the lower  
right corner of the front case.  
3. Place the oscilloscope face up on a soft surface (such as an anti-static mat),  
with the bottom facing you.  
4. Disconnect the bezel button flex circuit by gently pulling the flex circuit out  
of the connector.  
Display  
Follow these steps to remove the display module. You need to have previously  
removed the rear case, power supply assembly, I/O board, and front case  
assembly. Use a TORX T-15 screwdriver tip. See Figure 5--2 on page 5--6.  
1. Disconnect the display power cable from the main board and the backlight  
cable from the inverter board.  
2. Remove the 4 screws that hold the display to the front chassis.  
3. Gently lift the display up and off of the chassis.  
CAUTION. Be careful when removing and reinstalling the Display module cables.  
If the connectors have bent pins or are installed incorrectly; the Display may be  
destroyed.  
Front Panel  
Follow these steps to remove the front panel module. You need to have  
previously removed the front case assembly. Use a TORX T-15 screwdriver tip.  
See Figure 5--2 on page 5--6.  
1. Remove the 9 screws holding the front panel board to the chassis.  
2. Lift the front panel board off of the chassis.  
4-11  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
Troubleshooting  
This section contains information and procedures to help you isolate a defective  
module in a DPO4000 series oscilloscope.  
WARNING. Before doing this or any other procedure in this manual, read the  
General Safety Summary and Service Safety Summary found at the beginning of  
this manual. Also, to prevent possible injury to service personnel or damage to  
electrical components, read Preventing ESD on page 4--1.  
Adjustment After Repair  
If you replace the main board, you must adjust the instrument after repair. Refer  
Adjustment Procedures beginning on page 3--1 for information about adjustment.  
Required Tools and  
Equipment  
You need the following equipment to troubleshoot the instrument.  
Tools and Equipment  
Example  
DMM  
3.5 or greater digit DMM  
4-12  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
Troubleshooting Procedure  
Figures 4--3 through 4--5 are troubleshooting procedure flowcharts. Use them to  
troubleshoot an instrument failure. Begin with Figure 4--3.  
NOTE. Before performing the troubleshooting procedure, remove the instrument  
cabinet (see page 4--9 ).  
Power on the  
instrument.  
Does  
the instrument  
display the power-up  
screen?  
Perform the AC Power Supply  
troubleshooting procedure  
(Figure 4-4).  
No  
Yes  
Does  
the instrument pass  
all the  
Perform the Module Isolation  
troubleshooting procedure  
(Figure 4-5).  
No  
diagnostics?  
Yes  
Done.  
Figure 4-3: Primary troubleshooting procedure  
4-13  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
Continued from  
Figure 4-3.  
¯ Power off oscilloscope.  
¯ Disconnect AC power cord.  
¯ Remove instrument case.  
¯ Check that all cables are  
connected and fully seated into  
board connectors.  
No  
¯ Reconnect loose cables.  
¯ Return to top of Figure 4-3.  
Are all cables  
connected?  
Yes  
Connect AC power cord.  
¯ Measure for +12 V DC ( 0.3  
V) on the connector pins. (See  
figure at right)  
!WARNING!  
Be careful when measuring near  
the AC power supply, as there  
are line voltages present on  
exposed parts.  
Are there  
+12 volts on the  
No  
Replace the AC Power  
Supply module.  
connector  
pins ?  
Yes  
AC power supply OK. Return  
the DPO4000 oscilloscope to  
the nearest Tektronix Service  
Center for repair.  
Figure 4-4: AC power supply troubleshooting procedure  
4-14  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
From the oscilloscope  
power-up diagnostics  
message display.  
Does  
the power-up  
screen report any  
errors?  
No  
Oscilloscope is operating OK.  
Yes  
¯ Push UTILITY.  
¯ Push System to select  
Self Test.  
¯ Push OK Run Self Test.  
¯ Determine failure from  
Self Test Results.  
¯ Use table to select module to  
replace.  
Failed Test  
Most Likely Cause  
Acquisition  
CPU  
Display  
Front Panel  
I/O  
Main board  
Main board  
Main board  
Front Panel (or Main board)  
Comm module (or Main board)  
Main board  
ROM  
App Modules  
Front Panel (or App Keys)  
Figure 4-5: Module isolation troubleshooting procedure  
4-15  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
Unpacking and Repacking Instructions  
This section contains the information needed to unpack the oscilloscope and  
repack it for shipment or storage.  
Unpacking  
The oscilloscope and its standard accessories are carefully packed at the factory  
in a shipping carton. If, upon receipt, damage to the shipping carton is evident,  
notify the shipper. Tektronix, Inc. is not responsible for damage caused during  
shipping.  
If you have not already done so, carefully remove the oscilloscope and its  
accessories from the shipping carton and inspect them for damage. Save the  
shipping carton for repacking or storage.  
Repacking  
Use a corrugated cardboard shipping carton having a test strength of at least  
275 pounds (125 kg) and with an inside dimension at least six inches (15.25 cm)  
greater than the instrument dimensions.  
If the instrument is being shipped to a Tektronix Service Center, enclose the  
following information:  
H
H
H
H
H
The owner’s address  
Name and phone number of a contact person  
Type and serial number of the instrument  
Reason for returning  
A complete description of the service required  
Seal the shipping carton with an industrial stapler or strapping tape.  
Mark the address of the Tektronix Service Center and also your own return  
address on the shipping carton in two prominent locations.  
The oscilloscope should be stored in a clean, dry environment. The following  
environmental characteristics apply for both shipping and storage:  
Storage  
H
H
Temperature range: --20 °C to +60 °C  
Altitude: To 15,000 m  
See the DPO4000 Technical Reference for a complete listing of the environmen-  
tal characteristics.  
4-16  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Mechanical Parts List  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Mechanical Parts List  
This chapter contains a list of the replaceable modules for the DPO4000 series  
oscilloscope. Use this list to identify and order replacement parts.  
Parts Ordering Information  
Replacement parts are available through your local Tektronix field office or  
representative.  
Changes to Tektronix products are sometimes made to accommodate improved  
components as they become available and to give you the benefit of the latest  
improvements. Therefore, when ordering parts, it is important to include the  
following information in your order.  
H
H
H
H
Part number (see Part Number Revision Level below)  
Instrument type or model number  
Instrument serial number  
Instrument modification number, if applicable  
If you order a part that has been replaced with a different or improved part, your  
local Tektronix field office or representative will contact you concerning any  
change in part number.  
Change information, if any, is located at the rear of this manual.  
Tektronix part numbers contain two digits that show the revision level of the  
part. For some parts in this manual, you will find the letters XX in place of the  
revision level number.  
Part Number Revision  
Level  
Part Number Revision Level  
Revision Level May Show as XX  
670-7918-03  
670-7918-XX  
When you order parts, Tektronix will provide you with the most current part for  
your product type, serial number, and modification (if applicable). At the time of  
your order, Tektronix will determine the part number revision level needed for  
your product, based on the information you provide.  
5-1  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Mechanical Parts List  
Module Servicing  
Modules can be serviced by selecting one of the following three options. Contact  
your local Tektronix service center or representative for repair assistance.  
Module Exchange. In some cases you may exchange your module for a reman-  
ufactured module. These modules cost significantly less than new modules and  
meet the same factory specifications. For more information about the module  
exchange program, call 1-800-TEK-WIDE, extension 6630.  
Module Repair and Return. You may ship your module to us for repair, after which  
we will return it to you.  
New Modules. You may purchase replacement modules in the same way as other  
replacement parts.  
Using the Replaceable Parts List  
This section contains a list of the mechanical and/or electrical components that  
are replaceable for the instrument. Use this list to identify and order replacement  
parts. The following table describes each column in the parts list.  
Parts List Column Descriptions  
Column Column Name  
Description  
1
Figure & Index Number  
Items in this section are referenced by figure and index numbers to the exploded view  
illustrations that precede the list  
2
Tektronix Part Number  
Serial Number  
Use this part number when ordering replacement parts from Tektronix  
3 and 4  
Column three indicates the serial number at which the part was first effective. Column four  
indicates the serial number at which the part was discontinued. No entries indicates the part is  
good for all serial numbers  
5
6
Qty  
This indicates the quantity of parts used  
Name & Description  
An item name is separated from the description by a colon (:). Because of space limitations, an  
item name may sometimes appear as incomplete. Use the U.S. Federal Catalog handbook  
H6-1 for further item name identification  
7
8
Mfr. Code  
This indicates the code of the actual manufacturer of the part  
This indicates the actual manufacturer’s or vendor’s part number  
Mfr. Part Number  
Abbreviations  
Abbreviations conform to American National Standard ANSI Y1.1--1972.  
5-2  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Mechanical Parts List  
Mfr. Code to Manufacturer  
Cross Index  
The table titled Manufacturers Cross Index shows codes, names, and addresses  
of manufacturers or vendors of components listed in the parts list.  
Manufacturers cross index  
Mfr.  
code  
Manufacturer  
Address  
City, state, zip code  
06383  
06915  
0B445  
0KB01  
0KB05  
PANDUIT CORP  
17303 RIDGELAND AVE  
5825 N TRIPP AVE P.O. BOX 804238  
312 EAST MAIN STREET  
810 SE SHERMAN  
TINLEY PARK, IL 60477-3048  
CHICAGO, IL 60646  
RICHCO  
ELECTRI-CORD MFG CO INC  
STAUFFER SUPPLY CO  
NORTH STAR NAMEPLATE INC  
WESTFIELD, PA 16950  
PORTLAND, OR 97214-4657  
HILLSBORO, OR 97124-6474  
LIBERTY LAKE, WA 99019  
DELAWARE WATER GAP, PA 18327-0650  
5750 NE MOORE COURT  
23201 E APPLEWAY  
1WNW6 ACCRA-FAB INC  
30817  
LAIRD TECHNOLOGIES WORLD  
ONE SHIELDING WAY PO BOX 650  
HEADQUARTERS  
NEC ELECTRONICS INC  
CHOMERICS INC  
4T165  
57003  
7X318  
80009  
2880 SCOTT BLVD PO BOX 58062  
23839 S. BANNING BLVD  
SANTA CLARA, CA 95052-2554  
CARSON, CA 90745  
KASO PLASTICS INC  
TEKTRONIX INC  
5720-C NE 121ST AVE, STE 110  
VANCOUVER, WA 98682  
BEAVERTON, OR 97077-0001  
14150 SW KARL BRAUN DR  
PO BOX 500  
80640  
ARTESYN TECHNOLOGY POWER  
CONVERSION  
7 ELKIN ST  
SOUTH BOSTON, MA 02127-1615  
83285  
STEWARD INC  
1200 E 36TH ST~PO BOX 510  
10B VAN DYKE AVENUE  
10156 TORINO  
CHATTANOOGA, TN 37401-0510  
NEW BRUNSWICK, NJ 08901  
VAICENTALLO 62/456 ITALY  
HILLSBORO, OR 97124  
S3109  
FELLER U.S. CORPORATION  
PATELEC-CEM  
TK1373  
TK2548  
TK6121  
TK6159  
TK6253  
TK6433  
XEROX CORPORATION  
TUMBLER CORP  
7431 EVERGREEN PARKWAY  
4241 BUSINESS CENTER DR  
2900 NW 229TH AVE  
FREMONT, CA 94538  
RAD TECHNOLOGIES  
VOLEX INC  
HILLSBORO, OR 97124  
646 CARRIBEAN DR  
SUNNYVALE, CA 94089-1108  
I-SHENG ELECTRIC WIRE & PLASTIC  
(KUNSHAN) CO LTD.  
289 ZING YANG ZHONG RD  
TECHNICAL DEVELOPMENT ZONE JIANGSU,  
CHINA CN  
TK6547  
SOLECTRON MILPITAS INC  
1177 GIBRALTAR DRIVE SOLECTRON  
BUILDING 9  
MILPITAS, CA 95035  
TK6612  
TK6619  
HANACO MANUFACTURING COMPANY  
ACE MOLD (SHANGHAI) COMPANY LTD  
3965 SCHAEFER AVENUE  
CHINO, CA 91710  
318 NORTH ROUND-THE-CITY ROAD  
SHANGHAI COMPREHENSIVE INDUSTRIAL  
DEVELOPMENT ZONE, FENGXIAN SHANGHAI,  
CN  
TK6709  
SOLECTRON (SHANGHAI)  
798 YING LUN RO AD~WALGAOQIAO FTZ  
PUDONG, SHANGHAI, CN CN 200131  
5-3  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Mechanical Parts List  
Exploded Views  
Figures 5--1 through 5--5 on the following pages show the module-level  
exploded views of the DPO4000 series oscilloscope. The adjacent page is the list  
of components for that exploded view, indexed by the numbers in the figure.  
3
2
4
4
1
5
Figure 5-1: Exploded view, case  
5-4  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Mechanical Parts List  
Replaceable Parts List  
Fig. &  
index  
number  
Tektronix part Serial no. Serial no.  
discont’d  
Mfr.  
code  
number  
effective  
Qty Name & description  
Mfr. part number  
5-1  
-1  
Case  
200-4908-00  
202-0441-00  
211-1174-00  
1
1
4
COVER, FRONT, PROTECTIVE  
TK6710 200-4908-00  
80009 202-0441-00  
0KB01 211-1174-00  
-2  
CASE, REAR  
-3  
SCREW,MACHINE; W/HEAVY PATCH THREADLOCKING  
MATERIAL; 6-32 X 0.312 L,PNH,STL CAD PLT,T15  
-4  
-5  
335-1186-00  
348-1798-00  
1
2
MARKER,IDENT;LABEL,REAR PANEL  
FOOT, FRONT  
80009  
80009  
335-1186-00  
348-1798-00  
5-5  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Mechanical Parts List  
6
5
4
7
8
3
9
2
1
10  
15  
Flex Circuit  
J13  
11  
20  
13  
12  
13  
19  
14  
16  
18  
17  
Figure 5-2: Exploded view: front panel and display, 4-channel instruments  
5-6  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Mechanical Parts List  
Replaceable Parts List  
Fig. &  
index  
number  
Tektronix part Serial no. Serial no.  
discont’d  
Mfr.  
code  
number  
effective  
Qty Name & description  
Mfr. part number  
5-2  
-1  
Front Panel  
202-0432-00  
335-1187-00  
335-1191-00  
335-1193-00  
366-0860-01  
366-0859-01  
366-0861-01  
366-0865-00  
358-0883-00  
366-0867-00  
335-1467-00  
335-1466-00  
335-1465-00  
335-1464-00  
335-1463-00  
335-1462-00  
335-1461-00  
335-1460-00  
335-1459-00  
335-1458-00  
335-1183-00  
200-4418-00  
1
1
1
1
5
6
2
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
FRONT CASE ASSEMBLY  
80009  
80009  
80009  
80009  
80009  
80009  
80009  
80009  
202-0432-00  
335-1187-00  
335-1191-00  
335-1193-00  
366-0860-01  
366-0859-01  
366-0861-01  
366-0865-00  
-2  
LABEL. FRONT PANEL, MODEL DPO4054  
LABEL. FRONT PANEL, MODEL DPO4104  
LABEL. FRONT PANEL, MODEL DPO4034  
KNOB, .685 DIA, SOFT TOUCH  
-3  
-4  
-5  
-6  
-7  
-8  
-9  
KNOB, .470 DIA, SOFT TOUCH  
KNOB, .925 DIA, SOFT TOUCH  
KNOB, JOG  
BUSHING (JOG SHUTTLE KNOB)  
TK2565 358-0883-00  
KNOB, SHUTTLE  
80009  
80009  
80009  
80009  
80009  
80009  
80009  
80009  
80009  
80009  
80009  
80009  
7X318  
366-0867-00  
335-1467-00  
335-1466-00  
335-1465-00  
335-1464-00  
335-1463-00  
335-1462-00  
335-1461-00  
335-1460-00  
335-1459-00  
335-1458-00  
335-1183-00  
200-4418-00  
LABEL. FRONT PANEL, OVERLAY, RUSSIAN  
LABEL. FRONT PANEL, OVERLAY, KOREAN  
LABEL. FRONT PANEL, OVERLAY, STD CHINESE  
LABEL. FRONT PANEL, OVERLAY, SIM CHINESE  
LABEL. FRONT PANEL, OVERLAY, PORTUGUESE  
LABEL. FRONT PANEL, OVERLAY, JAPANESE  
LABEL. FRONT PANEL, OVERLAY, SPANISH  
LABEL. FRONT PANEL, OVERLAY, GERMAN  
LABEL. FRONT PANEL, OVERLAY, ITALIAN  
LABEL. FRONT PANEL, OVERLAY, FRENCH  
LABEL. FRONT PANEL, 4CH  
10  
11  
COVER; OPTION KEY DOOR,0.7270 X 1.2250 X 0.310,PC/  
ABS,FR110,SILVER GRAY  
12  
13  
260-2798-00  
211-1174-00  
1
9
SWITCH,KEYPAD;ELASTOMERIC FRONT PANEL  
80009  
260-2798-00  
SCREW,MACHINE; W/HEAVY PATCH THREADLOCKING  
MATERIAL; 6-32 X 0.312 L,PNH,STL CAD PLT,T15  
0KB01 211-1174-00  
14  
15  
16  
672-6132-01  
335-1194-00  
348-1860-00  
1
1
2
FRONT PANEL BOARD, 4 CH  
80009  
80009  
57003  
672-6132-00  
335-1194-00  
MARKER, FRONT PANEL, I/O  
GASKET, ELEC; CONDUCTIVE URETHANE FOAM  
77-03-3150-09600-  
AP  
17  
18  
19  
20  
441-2378-00  
650-4715-00  
174-5014-00  
211-1174-00  
1
1
1
4
CHASSIS;FRONT; SAFETY CONTROLLED  
ASSEMBLY, LCD DISPLAY  
TK6709 441-2378-00  
80009  
80009  
650-4715-00  
174-5014-00  
CABLE ASSY,SP;DISPLAY DATA,FLAT FLEX  
SCREW,MACHINE; W/HEAVY PATCH THREADLOCKING  
MATERIAL; 6-32 X 0.312 L,PNH,STL CAD PLT,T15  
0KB01 211-1174-00  
5-7  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Mechanical Parts List  
1
6
7
2
5
3
4
Figure 5-3: Exploded view: front panel 2-channel instruments  
5-8  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Mechanical Parts List  
Replaceable Parts List  
Fig. &  
index  
number  
Tektronix part Serial no. Serial no.  
number  
Mfr.  
code  
effective  
discont’d  
Qty Name & description  
Mfr. part number  
5-3  
-1  
-2  
-3  
-4  
-5  
-6  
-7  
Front Panel  
335-1189-00  
335-1185-00  
260-2812-00  
672-6133-01  
200-4906-00  
366-0860-01  
366-0859-01  
1
1
1
1
2
3
4
LABEL. FRONT PANEL, MODEL DPO4032  
80009  
80009  
335-1189-00  
335-1185-00  
MARKER,IDENT;LABEL,FRONT PANEL,2CH  
SWITCH,KEYPAD;ELASTOMERIC FRONT PANEL  
CIRCUIT BOARD; FRONT PANEL;2 CHANNEL  
COVER;BNC FILLER  
TK6612 260-2812-00  
80009 672-6133-01  
TK6433 200-4906-00  
KNOB, .685 DIA, SOFT TOUCH  
80009  
80009  
366-0860-01  
366-0859-01  
KNOB, .470 DIA, SOFT TOUCH  
5-9  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Mechanical Parts List  
2
1
3
2
10  
4
2
8
5
9
2
I/O board assemly  
J310  
7
Main board  
J780  
Main board  
J850  
6
Figure 5-4: Exploded view: power supply and I/O board  
5-10  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Mechanical Parts List  
Replaceable Parts List  
Fig. &  
index  
number  
Tektronix part Serial no. Serial no.  
discont’d  
Mfr.  
code  
number  
effective  
Qty Name & description  
Mfr. part number  
5-4  
-1  
Main Board  
407-5019-00  
211-0722-00  
211-1213-00  
119-7223-00  
1
BRACKET;POWER SUPPLY COVER  
TK6709 407-5019-00  
-2  
19  
1
SCREW,MACHINE; 6-32 X 0.250,PNH,STL,CDPL,T-15 TORX 0KB01 211-0722-00  
-3  
SCREW, METRIC 5 X 10MM, FLATHEAD, STL, ZINC, POSI  
0KB01 211-1213-00  
TK6159 119-7223-00  
-4  
1
FAN,TUBEAXIAL; 12VDC,0.09A,1.4W,19  
CFM,4000RPM,31DBA,60MM X 60MM X 25MM,W/7IN LEADS  
& CONNECTOR,TACH OUT  
-5  
-6  
-7  
-8  
-9  
10  
650-4939-00  
366-0844-00  
672-6131-01  
335-1213-00  
335-1214-00  
337-4378-00  
1
1
1
1
1
1
POWER SUPPLY SUB ASSEMBLY;DOMESTIC  
PUSH BUTTON;POWER; SAFETY CONTROLLED  
I/O BOARD  
80009  
650-4939-00  
TK6619 366-0844-00  
80009  
80009  
80009  
80009  
672-6131-01  
335-1213-00  
335-1214-00  
337-4378-00  
MARKER,IDENT;LABEL,REAR PANEL,LEFT  
MARKER,IDENT;LABEL,REAR PANEL, RIGHT  
SHIELD;INSULATOR,ELECTRICAL  
5-11  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Mechanical Parts List  
Display assembly  
to Inverter board  
assembly  
1
7
CN2  
2
Main board  
Inverter board  
assembly  
3
4
CN1  
J570  
J571  
5
6
9
Main board  
J930  
8
I/O board  
assemly  
J610  
Main board  
Figure 5-5: Exploded view: mainboard, inverter board, and side fans  
5-12  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Mechanical Parts List  
Replaceable Parts List  
Fig. &  
index  
number  
Tektronix part Serial no. Serial no.  
number  
Mfr.  
code  
effective  
discont’d  
Qty Name & description  
Mfr. part number  
5-5  
-1  
Main Board  
650-4937-00  
119-6806-00  
1
1
FAN TRAY ASSEMBLY  
80009  
4T165  
650-4937-00  
104PW191  
-2  
POWER SUPPLY; INVERTER,1200VAC NO-LOAD & 600VAC  
@ 85KOHM LOADED OUTPUT  
-3  
-4  
-5  
-6  
211-1143-00  
407-5113-00  
174-5015-00  
671-5843-00  
671-5959-00  
671-5960-00  
671-5961-00  
211-0722-00  
2
1
1
1
1
1
1
6
SCREW,PT; K35-1.57,PAN HEAD,T-15  
0KB01  
211-1143-00  
407-5113-00  
174-5015-00  
671-5843-00  
671-5959-00  
671-5960-00  
671-5961-00  
211-0722-00  
BRACKET,PCB;BACKLIGHT BOARD  
TK6433  
80009  
CABLE ASSY,SP;INVERTER POWER,FLAT FLEX  
CIRCUIT BD ASSY; MAIN BOARD, (DPO4104)  
CIRCUIT BD ASSY;350MHZ 4 CHANNEL, (DPO4054)  
CIRCUIT BD ASSY;350MHZ 2 CHANNEL, (DPO4034)  
CKT BD ASSY; 2 CHANNEL, (DPO4032)  
TK6547  
TK6547  
TK6547  
TK6547  
-7  
SCREW,MACHINE; 6-32 X 0.250,PNH,STL,CDPL,T-15 TORX 0KB01  
DR  
-8  
-9  
174-5014-00  
211-1174-00  
1
CABLE ASSY,SP;DISPLAY DATA,FLAT FLEX  
80009  
174-5014-00  
211-1174-00  
13  
SCREW,MACHINE; W/HEAVY PATCH THREADLOCKING  
MATERIAL; 6-32 X 0.312 L,PNH,STL CAD PLT,T15  
5-13  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Mechanical Parts List  
Replaceable Parts List  
Fig. &  
index  
number  
Tektronix  
part number  
Serial no. Serial no.  
effective discont’d  
Mfr.  
code  
Qty Name & description  
STANDARD ACCESSORIES  
A c c e s s o r ie s  
Mfr. part number  
- - -  
161-0104-05  
161-0104-06  
161-0104-07  
161-0167-00  
161-0306-00  
161-0066-09  
161-0298-00  
161-0066-00  
- - - - - - - -  
1
1
1
1
1
1
1
1
1
CA ASSY,PWR:3,1.0MM SQ,250V/10A,2.5 M  
(OPTION A3 - AUSTRALIAN)  
TK1373 161-0104-05  
CA ASSY,PWR:3,1.0MM SQ,250V/10A,2.5 M  
(OPTION A1 - EUROPEAN)  
S3109  
1923  
CA ASSY,PWR:3,1.0MM SQ,240V/10A,2.5 M  
(OPTION A2 - UNITED KINGDOM)  
TK6121 21072WH518.2  
CA ASSY,PWR:3,0.75MM SQ,250V/10A,2.5 M  
(OPTION A5 - SWITZERLAND)  
0B445  
161-0167-00  
CA ASSY,PWR:3,1.0MM SQ,250V/10A,2.5 M  
(OPTION A10 - CHINA)  
TK6253 92-2637-250BKH  
TK6121 2171H518.2  
CA ASSY,PWR:3,1.0MM SQ,250V/10A,2.5 M  
(OPTION A11 - INDIA)  
CA ASSY,PWR:3,1.0MM SQ,250V/10A,2.5 M  
(OPTION A6 - JAPAN)  
0B445  
S3109  
161-0298-00  
161-0066-00  
CABLE ASSY,PWER,:3,18 AWG,92 L  
(STANDARD CABLE)  
POUCH, PLASTIC:POUCH  
TK2582 TK1441 BLACK  
CORDURA  
5-14  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Index  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Index  
A
I
acquisition system, description, 2-2  
adjustment procedure, 3-2  
factory, 3-3  
I/O board removal, 4-10  
inspection  
exterior, 4-3  
adjustments, 3-1  
application modules, ix  
interior, 4-4  
inspection and cleaning  
exterior, 4-2  
flat panel display, 4-2  
interior, 4-2  
instrument calibration, 3-1  
internal modules, 4-8  
inverter board removal, 4-10  
I/O board connections, 2-2  
C
cabinet, 4-7  
calibration due date, 3-1  
calibration switch, 3-3  
circuit description, 2-1–2-4  
cleaning, 4-5  
cleaning and inspection  
exterior, 4-2, 4-3  
flat panel display, 4-2  
interior, 4-2  
K
keypad, 4-8  
connectivity, ix  
L
lubrication, 4-3  
D
description, circuit theory, 2-1–2-4  
diagram, 2-1  
display module removal, 4-11  
display system, description, 2-2  
documentation, ix  
M
main board functions, 2-2  
main board removal, 4-10  
manual conventions, ix  
manuals, ix  
module interconnections, 2-1  
E
electrostatic discharge, 4-1  
precautions, 4-1–4-2  
environmental characteristics, 4-16  
P
packing, 4-16  
power converter, description, 2-3  
power supply, 2-2  
power supply removal, 4-9  
preventive maintenance, 4-1  
probe compensation, 2-3  
procedures  
F
factory adjustment, 3-2  
fans removal, 4-10  
flat panel display, cleaning, 4-2  
flowcharts, troubleshooting, 4-13  
front case removal, 4-10  
front panel removal, 4-11  
front panel, description, 2-3  
flat panel cleaning, 4-2  
inspection and cleaning, 4-1–4-5  
interior cleaning, 4-5  
processor system, description, 2-3  
Index-1  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Index  
replaceable part, definition, ix  
R
rear case removal, 4-9  
removal procedures, 4-6, 4-8  
display module, 4-11  
fans, 4-10  
T
trigger system, description, 2-2  
trim, 4-7  
troubleshoot, 4-12  
front case, 4-10  
front panel, 4-11  
I/O board, 4-10  
inverter board, 4-10  
main board, 4-10  
power supply, 4-9  
rear case, 4-9  
U
user manual, 1-1  
Index-2  
DPO4000 Series Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Techno Source Handheld Game System Titanium Edition User Manual
Tiger Games 44900 User Manual
Toshiba CRT Television 32AF62 User Manual
Tripp Lite Indoor Furnishings SRCABLELADDER User Manual
Uniden Stereo System UBC100XLT User Manual
Uniden Two Way Radio LTD 715 User Manual
Universal Remote Control Universal Remote omega MX 650 User Manual
ViewSonic MP3 Player VPD500 User Manual
Viking Convection Oven Built In Convection Microwave Hood User Manual
Viking Electronics Cooktop F20033E User Manual