Seiko Instruments Heat Pump MT 17E 003 D User Guide

MT-17E-003-D  
STP-H600/H1000 Series  
Turbomolecular Pump  
INSTRUCTION MANUAL  
(Third Edition-d)  
Read through the Safety Precautions of this Manual  
carefully before using the STP pump.  
Keep this Manual in a place where you can quickly access  
it at any time.  
Seiko Instruments Inc.  
Dec. 2001  
Copyright 2001 Seiko Instruments Inc. All rights reserved. Printed in Japan.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
WARNING  
Check the properties of the gas to be used, referring to the Material Safety Data Sheet  
(MSDS) you obtain from the gas supplier. Follow any and all safety precautions and/or  
recommendations instructed by the gas supplier.  
Take appropriate measures specified in the MSDS to prevent a problem if you use any  
corrosive, reactive, flammable or other system gas. Dilute the pumped gas with a monitored  
inert gas if necessary. Take appropriate measures so as not to cause a problem due to the  
pumped gas  
The STP pump is provided with a high-speed rotor. Secure the STP pump according to the  
specified method. Failure to do so may lead to serious personal injury, product and/or  
peripheral equipment damage if any abnormality/error occurs in the rotor.  
The STP pump operates at high temperatures while the baking heater is in operation.  
NEVER touch the STP pump and its peripheral equipment while the baking heater is in  
operation.  
Operators can burn hands.  
Always exhaust residual gases thoroughly from the STP pump when removing the STP  
pump from the vacuum equipment. Residual gases in the STP pump may cause an  
accident which, for certain gases, may involve serious injury or death.  
Confirm the characteristics of the gas to be used, referring to the Material Safety Data Sheet  
(MSDS) you obtain from the gas supplier. Wear personal protective equipment if  
necessary.  
When covering the battery case, be sure not to catch the battery cable in the battery case.  
Failure to do so may short-circuit the battery and result in a leakage, production of fire or  
explosion.  
NEVER throw the battery into fire nor heat it.  
Failure to do so may result in an explosion, production of fire and/or serious personal injury.  
Observe the instructions given by the battery maker, the national and/or local government  
when disposing of the battery.  
Exhaust residual gas thoroughly when disposing of the STP pump. If the STP pump is  
used for any toxic or reactive gas, always clean the STP pump and dispose of it as industrial  
waste in accordance with guidelines given by the national and/or local government.  
Residual gas in the STP pump may cause an accident which, for certain gases, may involve  
serious injury or death.  
Always remove the battery from the STP control unit when disposing of it. Failure to do so  
may result in fire or other accident.  
2
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
CAUTION  
NEVER use any gas that is not specified as usable in this Manual. The use of such gas  
may corrode the STP pump and damage it.  
A hazardous live voltage may exist at connector/terminal that marked  
.
DO NOT touch the terminal. Doing so may result in electric shock.  
When operating connection/disconnection to terminal, always power OFF the STP pump  
(Switch the breaker “OFF”).  
Always check the STP pump has stopped, then turn OFF the primary power (switch the  
breaker “OFF”) before proceeding to any of the following operations. Failure to do so may  
cause the STP pump to rotate accidentally, which may injure operators seriously or result in  
electric shock.  
Sꢀ Connect or disconnect cables;  
Sꢀ Connect an external battery;  
Sꢀ Perform maintenance and inspections such as replacement of the internal battery  
and/or fuses as well as inspections of deposit and/or the air cooling fan (when replacing  
fuses, always disconnect the internal battery connector);  
Sꢀ Perform investigations into probable causes and action/measures taken in the event of  
occurrence of a problem; or  
Sꢀ Open and close the STP control unit front panel (hereinafter referred to as the front  
panel) in order to check or reset the "BATTERY NG" (hereinafter referred to as  
"BATTERY NG") lamp.  
The STP pump and the STP control unit are heavy products. Always use a crane or the  
like when lifting them. When lifting the STP pump by hands for unavoidable reasons, it  
must always be lifted by two or more people. Failure to do so may damage their hipbone or  
injure them due to occurrence of an accident such as fall.  
Install the STP control unit not only by fitting it with the front panel fitting screws but also by  
supporting it from the bottom side. Fitting the STP control unit with the screws only cannot  
sustain its weight, and therefore resulting in product damage.  
Always use the STP pump, STP control unit and STP connection cables of the same model  
name, serial number and cable length, which are specified on their own name plate.  
Failure to do so may result in product damage. If you use units with the same model name  
but different serial number and cable length; they must be adjusted. If the model names  
are different, they may not be used even when performing adjustment. In both cases,  
contact Seiko Instruments.  
Use the STP connection cable and the motor connection cable that have a label affixed  
STP-H600/H1000  
Series  
.
The use of different cables may result in product damage.  
Connect the cables securely. NEVER bend nor place heavy objects on the cable. Doing  
so may result in electric shock or product damage.  
Always use the power voltage specified on the name plate for the primary power voltage of  
the STP control unit. Wire the power cable securely. Incorrect wiring may result in electric  
shock or product damage.  
NEVER remove the splinter shield from the STP pump. Doing so may result in product  
damage.  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
CAUTION  
DO NOT put foreign objects into the STP pump. Doing so may result in product damage.  
NEVER turn OFF the primary power (DO NOT switch the breaker “OFF”) while the STP  
pump is rotating. Doing so may result in product damage.  
Perform investigations into probable causes and remove them before restarting the STP  
pump in the event of occurrence of a problem. The use of the abnormal STP pump may  
result in product damage.  
Always replace batteries once a year. Failure to do so, the battery backup operation may  
not run at the power failure.  
DO NOT connect internal batteries and external batteries simultaneously. Failure to do so  
may result in product damage.  
DO NOT move the STP pump and the STP control unit while the STP pump is in operation.  
Doing so may result in product damage.  
4
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
INTRODUCTION  
Thank you very much for purchasing Seiko Instruments’s turbomolecular pump. The  
turbomolecular pump is designed to be installed in the vacuum equipment in to exhaust  
gases from it.  
This manual covers all items necessary to ensure safe installation, operation and  
maintenance of the following series of the STP-H600/H1000 turbomolecular pump:  
Model Name  
Specification  
Sꢀ STP-H600  
Sꢀ STP-H1000  
Sꢀ STP-H600C  
Sꢀ STP-H1000C  
High-throughput type  
High-throughput type  
High-throughput type, chemical specific *1  
High-throughput type, chemical specific  
For the specifications of other models of the STP-H600/H1000 pump series, contact Seiko  
Instruments.  
In this manual, the above STP pump series is collectively referred to as the “STP pump.”  
APPLIED STANDARDS  
The STP pump conforms to the following directives and standards:  
Applied Directives  
Sꢀ EC Machinery Directive  
Sꢀ EC Electromagnetic Compatibility Directive  
Sꢀ EC Low Voltage Directive  
Applied Standards  
Sꢀ EN292-1  
Sꢀ EN292-2  
Sꢀ EN60204-1  
Sꢀ EN55011 (class A)  
Sꢀ EN50082-2  
Applied Standards (Only a some model)*2  
Sꢀ UL3101-1,1993 (Electrical Equipment for Laboratory Use; Part 1 : General  
Requirements)  
*1 Chemical specific: STP pump with anti-corrosive treatment (responding to chlorine, fluorine or other system gases)  
*2 Contact Seiko Instruments about recognized model.  
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
STP-H600/H1000 Series Instruction Manual  
PRECAUTIONS  
1) No part of this manual may be reproduced in any form by any means  
written permission from Seiko Instruments.  
without prior  
2) Seiko Instruments pursues a policy of continuing improvement in design and  
performance of this product. The right is, therefore, reserved to vary specifications and  
design without notice. Understand that the product you purchased and its contents  
including specifications described in this manual may differ.  
REQUEST  
If you find inaccuracies or errors in this manual, advise your nearest sales representative  
office of Seiko Instruments or the following sections of them:  
Seiko Instruments Inc.  
Sales Office (Japan)  
Seiko Instruments Inc. Vacuum Innovative Products Sales Department  
1-8 Nakase, Mihama-ku, Chiba-shi, Chiba 261-8507 JAPAN  
Telephone:  
<Vacuum Innovative Products Sales Department, Direct>  
Domestic  
043-211-1401  
International  
Domestic  
+ 81-43-211-1401  
043-211-8075  
Facsimile:  
International  
+ 81-43-211-8075  
Manufacturer, Sales Office (Japan)  
Seiko Instruments Inc. Vacuum Innovative Products Division  
4-3-1 Yashiki, Narashino-shi, Chiba 275-0004 JAPAN  
Telephone:  
<Vacuum Innovative Products Service Section, Direct>  
Domestic  
047-475-3123  
International  
<Company main>  
Domestic  
+81-47-475-3123  
047-475-3111  
International  
Domestic  
International  
+81-47-475-3111  
047-471-1544  
+81-47-471-1544  
Facsimile:  
Copyright 2001 Seiko Instruments Inc. All rights reserved.  
Vacuum Innovative Products Sales Department  
1-8 Nakase, Mihama-ku, Chiba 261, Japan  
Telephone:  
Facsimile:  
+ 81-43-211-1401(Direct)  
+ 81-43-211-8075  
6
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
LIMITED WARRANTY  
This WARRANTY applies to the customer to whom Seiko Instruments has delivered this  
product.  
1. WARRANTY PERIOD:  
Seiko Instruments warrants this product against defects for a period of  
two (2) years from the date of delivery or during the period specified in  
the agreement made by and between the customer and Seiko  
Instruments.  
2. ITEM WARRANTED:  
1) This warranty applies only to the product delivered from Seiko  
Instruments to the customer.  
2) If any defect is found during this period, Seiko Instruments will, at  
its option, repair or recondition the product free of charge. The  
costs for repair or replacement of the product after the warranty  
period has passed will be at the customer's own charge.  
3. DISCLAIMER:  
Seiko Instruments makes no warranty with respect to any damage occurred  
due to any of the following during the warranty period:  
1) Handling, operation or maintenance other than that specified herein;  
2) Failure to follow any of the warnings or cautions enumerated under  
or  
;
3) Installation, operation or maintenance using parts which are not  
specified by Seiko Instruments;  
4) Maintenance personnel other than those authorized by Seiko  
Instruments or its specified plant have disassembled, reconditioned, or  
tampered the product;  
5) Defect resulting from the not-specified use of the product;  
6) When the product is used under special conditions without obtaining  
the written consent of Seiko Instruments (Particular gases, strong  
magnetic field and the radiation are added to the product.);  
7) Defect resulting from deposit;  
8) Water cooling system defect resulting from water quality used;  
9) Defect resulting from the installation of the product (Exclude the  
installation by authorized personnel.);  
10) Deterioration in the external because of use (Discoloration, scratches  
and so forth);  
11) Product damage occurred during transport or other factors not  
attributable to Seiko Instruments;  
12) Product breakage or damage due to natural disasters, fire or other  
external factors;  
13) Deterioration in the basic performance due to the use of the product  
beyond limits of the use;  
14) Any direct, incidental or consequential damage resulting from the use  
of the product;  
15) When continuously operated without overhaul after the WARNING  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
indication (“WARNING” message) on the LCD display;  
16) Overhaul and replacement of maintenance parts;  
4. SPARE PARTS:  
Fuse and air cooling fan for control unit  
Touch down bearing  
Heater  
8
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
TABLE OF CONTENTS  
SAFETY PRECAUTION  
INTRODUCTION  
LIMITED WARRANTY  
1 Precautions for Safe Operation of the STP Pump..................................... 1-1  
1.1  
1.2  
1.3  
1.4  
Usable Gases............................................................................................................ 1-1  
Maintenance Precautions.......................................................................................... 1-1  
Inspection Precautions.............................................................................................. 1-2  
Labels........................................................................................................................ 1-2  
2 Unpacking.................................................................................................. 2-1  
2.1  
2.2  
Unpacking the STP Pump......................................................................................... 2-1  
Unpacking the STP Control Unit ............................................................................... 2-2  
3 Installation of the STP pump...................................................................... 3-1  
3.1  
3.2  
Name and Function of Each Part.............................................................................. 3-1  
Precautions Before Installation ................................................................................. 3-3  
3.2.1 Operating Environment ................................................................................ 3-3  
3.2.2 Installation Area ........................................................................................... 3-4  
3.2.3 Bench........................................................................................................... 3-4  
How to Install the STP Pump .................................................................................... 3-5  
3.3.1 Cleaning the Seal......................................................................................... 3-6  
3.3.2 STP Pump Installation Positions.................................................................. 3-7  
3.3.3 How to Secure the STP Pump ..................................................................... 3-8  
3.3.4 Vacuum Piping........................................................................................... 3-12  
3.3.5 Connecting the Emergency Vent. Valve .................................................... 3-14  
3.3.6 Connecting the Purge Port......................................................................... 3-14  
3.3.7 Connecting the Ground Cable.................................................................... 3-15  
3.3  
4 Installation of the STP Control Unit............................................................ 4-1  
4.1  
Name and Function of Each Part.............................................................................. 4-1  
4.1.1 Front Panel................................................................................................... 4-1  
4.1.2 Rear Panel ................................................................................................... 4-4  
4.1.3 Inside of the STP Control Unit...................................................................... 4-6  
Precautions Before Installation ................................................................................. 4-8  
4.2.1 Operating Environment ................................................................................ 4-8  
4.2.2 Installation Area ........................................................................................... 4-9  
Attaching the STP Control Unit Front Panel to a Rack ........................................... 4-10  
Cable Connection.................................................................................................... 4-11  
4.4.1 Name and Dimensions of Each Cable....................................................... 4-11  
4.4.2 How to Connect the Cables ....................................................................... 4-13  
4.2  
4.3  
4.4  
5 How to Start/Stop the STP Pump .............................................................. 5-1  
5.1  
5.2  
5.3  
5.4  
5.5  
Before Starting .......................................................................................................... 5-1  
Starting/Stopping Time.............................................................................................. 5-1  
Start Procedures ....................................................................................................... 5-2  
Stop Procedures ....................................................................................................... 5-2  
Manual Operation...................................................................................................... 5-3  
5.5.1 Powering ON................................................................................................ 5-3  
5.5.2 Starting the STP Pump ................................................................................ 5-3  
5.5.3 Stopping the STP Pump............................................................................... 5-3  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
5.5.4 Starting the STP Pump after Stopping......................................................... 5-3  
5.5.5 Powering OFF.............................................................................................. 5-4  
Remote Operation..................................................................................................... 5-5  
5.6.1 Powering ON................................................................................................ 5-5  
5.6.2 Starting/Stopping the STP Pump ................................................................. 5-5  
5.6.3 Starting the STP Pump After Stopping......................................................... 5-5  
5.6.4 Powering OFF.............................................................................................. 5-6  
5.6  
6 Safety Functions When an Abnormality/Error Occurs ............................... 6-1  
6.1  
Safety Functions........................................................................................................ 6-1  
6.1.1 Power Failure............................................................................................... 6-1  
6.1.2 Abnormal State of Magnetic Bearing ........................................................... 6-3  
6.1.3 Excessive Vibration...................................................................................... 6-3  
6.1.4 Inverter Overload ......................................................................................... 6-3  
6.1.5 Overheating Inside the STP pump............................................................... 6-3  
6.1.6 Overheating Inside the STP Control Unit..................................................... 6-3  
6.1.7 Overspeed.................................................................................................... 6-3  
6.1.8 Abnormal Battery Voltage ............................................................................ 6-4  
6.1.9 Failure of Brake............................................................................................ 6-4  
Restarting after Any Safety Function Operates......................................................... 6-5  
Operation of the Emergency Vent. Valve.................................................................. 6-7  
6.2  
6.3  
7 Gas Suction, Cooling and Baking the STP Pump...................................... 7-1  
7.1  
Gas Suction............................................................................................................... 7-1  
7.1.1 How to Introduce a Purge Gas..................................................................... 7-1  
Cooling the STP Pump.............................................................................................. 7-2  
7.2.1 Water Cooling Method ................................................................................. 7-2  
7.2.2 Air Cooling Method....................................................................................... 7-3  
Baking the STP Pump............................................................................................... 7-4  
7.3.1 Attaching a Baking Heater ........................................................................... 7-4  
7.3.2 Interlocking the STP Pump Operation with Baking.......................................... 7-6  
7.2  
7.3  
8 Remote Input/Output Signal Terminal Blocks............................................ 8-1  
8.1  
8.2  
8.3  
8.4  
I/O TB2 Terminal Block............................................................................................. 8-1  
DC I/O TB3 Terminal Block....................................................................................... 8-6  
START STOP TB5 Terminal Block........................................................................... 8-8  
START/STOP TB6 Terminal Block........................................................................... 8-9  
9 Internal Battery........................................................................................... 9-1  
9.1  
9.2  
9.3  
9.4  
9.5  
Life of the Internal Battery......................................................................................... 9-1  
Allowable Shelf Life of the Internal Battery................................................................ 9-1  
How to Charge the Internal Battery........................................................................... 9-2  
BATTERY NG Lamp ................................................................................................. 9-3  
How to Replace the Internal Battery.......................................................................... 9-4  
9.5.1 How to Replace the Battery Case ................................................................ 9-4  
9.5.2 How to Replace the Battery in the Battery Case.......................................... 9-6  
How to Dispose of the Internal Battery...................................................................... 9-8  
9.6  
10 External Battery ....................................................................................... 10-1  
10.1 Specifications for the External Battery .................................................................... 10-1  
10.2 Installation of the External Battery........................................................................... 10-2  
10.3 How to Charge the External Battery........................................................................ 10-3  
10.4 How to Dispose of the External Battery................................................................... 10-3  
11 Operation Principle of the STP Pump...................................................... 11-1  
10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
12 Maintenance and Inspection.................................................................... 12-1  
12.1 Replacing the Internal Battery................................................................................. 12-1  
12.2 Inspecting the Air Cooling Fan................................................................................ 12-1  
12.3 Replacing the Fuses ............................................................................................... 12-2  
12.4 Inspecting for Deposit ............................................................................................. 12-3  
12.5 Overhaul.................................................................................................................. 12-4  
12.6 Transporting for Repair or Overhaul ....................................................................... 12-5  
13 Storage .................................................................................................... 13-1  
13.1 The STP Pump........................................................................................................ 13-1  
13.2 The STP Control Unit.............................................................................................. 13-2  
13.3 Restarting Precautions............................................................................................ 13-2  
14 Disposal ................................................................................................... 14-1  
14.1 The STP Pump........................................................................................................ 14-1  
14.2 The STP Control Unit.............................................................................................. 14-1  
14.3 The Battery.............................................................................................................. 14-2  
15 Troubleshooting....................................................................................... 15-1  
15.1 Troubleshooting Immediately After An Abnormality/Error Occurs .......................... 15-1  
15.2 Abnormalities When Powering ON ......................................................................... 15-2  
15.3 Abnormalities When Performing The STP Pump Start Operation.......................... 15-3  
15.4 Abnormalities While the STP Pump is Rotating...................................................... 15-3  
15.5 When Any Abnormality/Error Warning Lamp Lights ............................................... 15-5  
16 Specifications and Accessories ............................................................... 16-1  
16.1 Specifications for the STP Pump ............................................................................ 16-1  
16.2 Specifications for the STP Control Unit................................................................... 16-2  
16.3 Accessories............................................................................................................. 16-4  
16.4 Recommended Spare Parts.................................................................................... 16-4  
STP PUMP PROBLEM CHECK SHEET  
UNIT CONVERSION TABLE  
11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
TABLES  
Table 3.1  
Table 3.2  
Tightening torque of bolt....................................................................................... 3-8  
Maximum Torque predicted and  
Recommended securing bolt for inlet port flange................................................. 3-9  
Number of Claw Clamps by Size of Flange........................................................ 3-10  
Connecting Primary Power Cable....................................................................... 4-15  
Starting/Stopping the STP Pump During Remote Operation................................ 5-5  
State of Lamps and REMOTE Output Signal at Power Failure ............................ 6-2  
Operation of the STP Pump after Power Recovery.............................................. 6-2  
Safety Functions ................................................................................................... 6-8  
Attaching Positions of the Cooling Unit and Baking Heater.................................. 7-5  
I/O TB2 Terminal Block ........................................................................................ 8-1  
Rated Contacts for Relays CR1 to 4, CR7, CR8 and CR10................................. 8-4  
Rated Contacts for Relays CR5 and 6.................................................................. 8-4  
DC I/O TB3 Terminal Block .................................................................................. 8-6  
START STOP TB5 Terminal Block....................................................................... 8-8  
START/STOP TB6 Terminal Block....................................................................... 8-9  
Table 3.3  
Table 4.1  
Table 5.1  
Table 6.1  
Table 6.2  
Table 6.3  
Table 7.1  
Table 8.1  
Table 8.2  
Table 8.3  
Table 8.4  
Table 8.5  
Table 8.6  
Table 10.1 Specifications for the External Battery................................................................ 10-1  
Table 15.1 Troubleshooting After Powering ON................................................................... 15-2  
Table 15.2 Troubleshooting When Performing The STP Pump Start Operation.................. 15-3  
Table 15.3 Troubleshooting While the STP Pump Is Rotating ............................................. 15-3  
Table 15.4 Cross Reference of Items of Abnormality Warning Lamps ................................ 15-5  
Table 15.5 Troubleshooting When Any of Abnormality/Error Warning  
Lamp Lights ........................................................................................................ 15-6  
Table 16.1 Specifications for the STP Pump........................................................................ 16-1  
Table 16.2 Specifications for the STP Control Unit .............................................................. 16-2  
Table 16.3 Accessories ........................................................................................................ 16-4  
Table 16.4 Recommended Spare Parts ............................................................................... 16-4  
12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
FIGURES  
Figure 2.1  
Figure 3.1  
Figure 3.2  
Figure 3.3  
Figure 3.4  
Figure 3.5  
Figure 3.6  
Example of Lifting the STP Pump......................................................................... 2-1  
Configuration of the STP Pump............................................................................ 3-2  
Installation of the STP Pump to the Vacuum Equipment...................................... 3-5  
STP Pump Installation Positions........................................................................... 3-7  
Positions of the Outlet Port on the Horizontally or Slanted Installed STP Pump .. 3-7  
Example of securing the STP pump (When securing the inlet port with bolts)..... 3-9  
Example of securing the STP pump  
(When securing the inlet port flange with claw clamps)...................................... 3-10  
Example of securing the STP pump  
Figure 3.7  
(When installing the damper in the inlet port flange) .......................................... 3-11  
Connecting the Purge Port and Emergency Vent. Valve.................................... 3-15  
STP Control Unit Front Panel ............................................................................... 4-3  
STP Control Unit Rear Panel................................................................................ 4-5  
Inside of the STP Control Unit .............................................................................. 4-7  
Peripheral Space of the STP Control Unit ............................................................ 4-9  
Example of Securing the STP Control Unit......................................................... 4-10  
External Dimensions of Each Cable ................................................................... 4-11  
External Dimensions of he Power Cable ............................................................ 4-12  
How to Secure Primary Power Cable ................................................................. 4-16  
I/O TB2 Remote Output Signal Terminal Block.................................................... 8-3  
Example of Connecting I/O TB2 R. Pump Terminal............................................. 8-5  
Example of Securing the Remote Cable............................................................... 8-5  
DC I/O TB3 Remote Input Signal Terminal Block................................................. 8-7  
START STOP TB5 Remote Input Signal Terminal Block..................................... 8-8  
START/STOP TB6 Remote Input Signal Terminal Block..................................... 8-9  
Life of the Internal Battery..................................................................................... 9-2  
Allowable Shelf Life of the Internal Battery ........................................................... 9-2  
How to Replace the Internal Battery ..................................................................... 9-7  
Figure 3.8  
Figure 4.1  
Figure 4.2  
Figure 4.3  
Figure 4.4  
Figure 4.5  
Figure 4.6  
Figure 4.7  
Figure 4.8  
Figure 8.1  
Figure 8.2  
Figure 8.3  
Figure 8.4  
Figure 8.5  
Figure 8.6  
Figure 9.1  
Figure 9.2  
Figure 9.1  
Figure 11.1 Cross Sectional View of the STP Pump ............................................................. 11-2  
Figure 16.1 External Appearance of the STP Pump  
(Example: STP-H1000 Series) ........................................................................... 16-5  
Figure 16.2 External Appearance of the STP Control Unit .................................................... 16-6  
Figure 16.3 Label Affixing Positions for the STP Pump......................................................... 16-7  
Figure 16.4 Label Affixing Positions for the STP Control Unit ............................................... 16-8  
Figure 16.4 Label Affixing Positions for the Special Accessory............................................. 16-9  
13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
1 Precautions for Safe Operation of the STP Pump  
1.1 Usable Gases  
Chlorine or fluorine system gases can be used in chemical specific pumps (STP-  
H600C/H1000C or other models). When you use gases including alkaline metals, but  
excluding Li, gases including Ga, Hg, In, or Sn, or HBr, contact Seiko Seiki.  
WARNING  
Check the properties of the gas to be used, referring to the Material Safety Data  
Sheet (MSDS) you obtain from the gas supplier. Follow any and all safety  
precautions and/or recommendations instructed by the gas supplier.  
Take appropriate measures specified in the MSDS to prevent a problem if you  
use any corrosive, reactive, flammable or other system gas. Dilute the pumped  
gas with a monitored inert gas if necessary. Take appropriate measures so as  
not to cause a problem due to the pumped gas  
CAUTION  
NEVER use corrosive gases (chlorine, fluorine, or other system gases) in the  
STP-H600/STP-H1000 pump or other models without anti-corrosion treatment.  
Introduce a dry N2 gas (purge gas) to protect the inside of the STP pump when  
using reactive or corrosive gases.  
Cool the STP pump to within the operating pressure range to prevent the STP  
pump from overheating when sucking gases.  
1.2 Maintenance Precautions  
Perform any maintenance of the STP pump and the STP control unit (battery replacement,  
fuse replacement), following Section 12, "Maintenance and Inspection".  
CAUTION  
Always turn OFF the primary power (switch the breaker "OFF") before  
performing any maintenance.  
Disconnect internal battery connector before performing maintenance of the STP  
control unit.  
NEVER touch any portions other than those designated when performing  
maintenance.  
Careless touch may cause electric shock and/or a short-circuiting of the internal  
circuit, resulting in product damage or a problem.  
1-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
1.3 Inspection Precautions  
Perform inspections of the STP pump and the STP control unit (check of the air cooling fan  
and reset of the "BATTERY NG" lamp), following Section 12, "Maintenance and Inspection"  
When performing inspections (check of the air cooling fan and reset of the "BATTERY NG"  
lamp), always turn OFF the primary power (switch the breaker "OFF") before opening the  
front panel, then turn ON the primary power again (switch the breaker "ON").  
CAUTION  
NEVER touch any portions other than those designated when performing  
inspections.  
Careless touch may cause electric shock and/or a short-circuiting of the internal  
circuit, resulting in product damage or a problem.  
1.4 Labels  
The following labels are affixed to the STP pump and STP control unit.  
Read the contents of the labels before operation. For the positions of the labels, see Figures  
Label Affixing Positions.  
(1) STP Pump Caution Label  
This label describes precautions for operating the STP pump.  
Follow these precautions.  
(2) STP Control Unit Caution Label  
This label describes precautions for operating the STP control unit.  
Follow these precautions.  
1-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
(3) Heavy Product Caution Label  
This label is affixed to the product with a mass of 20 kg or more.  
Follow the precautions of Section 2, "Unpacking" so as not to cause any accident during  
handling.  
(4) STP Pump Installation Warning Label  
This label describes installation of the STP pump.  
Install the STP pump according to the precautions of Section 3, "Installation of the STP  
Pump. "  
(5) Emergency Vent. Valve Caution Label  
This label describes the emergency vent. valve installation position.  
Connect the emergency vent. valve before use.  
1-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
(6) Connector Caution Label  
This label describes lock of the connector.  
This label instructs operators to prevent the connectors from being disconnected while  
the STP pump is in operation.  
(7) STP Control Unit Safety Instruction Label  
This label describes instructions before operating the STP control unit.  
(8) High Voltage Device Caution Label  
The STP control unit is equipped with a high voltage device. This label warns operators to  
pay attention to the high voltage device.  
1-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
(9) Battery Instruction Label  
This label instructs operators to replace batteries once a year.  
The next replacement date of batteries is specified upon delivery of the STP pump.  
Record the next replacement date (after one year) of batteries in the blank of the label when  
replacing them.  
This label describes precautions for use of external batteries.  
(10) Rotational Direction Instruction Label  
This label describes the rotational direction of the STP pump.  
The STP pump rotates in this direction.  
ROTATION  
(11) Voltage Rating Label  
This label describes the rated voltage of the STP control unit.  
Use voltage specified in this label.  
200 240V  
1-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
(12) Hot Surface Warning Label  
This label instructs operators so as not to touch the hot surface of the STP pump.  
The use of the baking heater (optional accessory) may lead to a considerable rise in  
temperatures outside the STP pump.  
This label warns operators so as not to burn hands.  
(only when using the baking heater.)  
(13) Tuning Caution Label  
This label describes precautions for performing the tuning.  
Tuning should be performed by Seiko Seiki trained or authorized personnel only.  
! 注意  
浮上調整を行う場合は弊社  
のトレーニングまたは許可を  
受けてください。  
!
CAUTION  
Tuning must be performed  
by Seiko Instruments Inc. or  
authorized personnel only.  
1-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
2 Unpacking  
2.1 Unpacking the STP Pump  
Check the following before unpacking the STP pump.  
1) Check the package for bruises, breakage, wetness, etc.  
If there is any abnormality/error or it is judged necessary to return the product, contact  
Seiko Seiki.  
2) Check the contents of the package.  
See Section 16.3, "Accessories."  
CAUTION  
The mass of the STP pump is approx. 30 kg. Use a crane or other appropriate  
means to lift the STP pump.  
Lift the STP pump with an eyebolt or a similar tool that uses the fitting hole  
attached to the inlet port flange.  
Lift it with the aid of a suspension tool under the base when installing the STP  
pump without fitting hole.  
Observe national laws/regulations, safety standards and so on when lifting the  
STP pump.  
Use a crane or other appropriate means sufficient enough to withstand the load  
when lifting the STP pump.  
Always lift the STP pump in stable places free of excessive shock or vibration to  
prevent it from shaking or dropping.  
Have at least two people lift the STP pump when doing so by hand.  
NOTICE  
Be careful not to scratch the flange of the STP pump.  
Before installing the STP pump, check whether or not there are scratches on the  
surface.  
It is recommended to keep the packaging materials, such as the corrugated  
fiberboard container and cushioning material for possible reuse.  
Figure 2.1 Example of Lifting the STP Pump  
2-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
2.2 Unpacking the STP Control Unit  
Check the following before unpacking the STP control unit.  
1) Check the package for bruises, breakage, wetness, etc.  
If there is any abnormality/error or it is judged necessary to return the product, contact  
Seiko Seiki.  
2) Check the contents of the package.  
See Section 16.3, "Accessories."  
CAUTION  
The mass of the STP control unit is approx. 40 kg. Use a crane or appropriate  
means to lift the STP control unit.  
Lift the STP control unit using the two handles attached to the front panel.  
Observe national laws/regulations, safety standards and so on when lifting the  
STP control unit.  
Use a crane or appropriate means sufficient enough to withstand the load when  
lifting the STP control unit.  
Always lift the STP control unit in stable places free of excessive shock or  
vibration to prevent it from shaking or dropping.  
Have at least two people lift the STP control unit when doing so by hand.  
NOTICE  
It is recommended to keep the packaging materials, such as the corrugated  
fiberboard container and cushioning material for possible reuse.  
2-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
3 Installation of the STP pump  
3.1 Name and Function of Each Part  
(1) Inlet Port Flange (ICF*1 , VG*2 , ASA, ISO, etc.)  
ꢀꢁ Connected to the vacuum equipment (at the high vacuum side).  
CAUTION  
A splinter shield is attached to the inlet port flange to prevent foreign particles  
from falling into the STP pump.  
NEVER remove it.  
(2) Outlet Port Flange (KF*2 40)  
ꢀꢁ Connected to the inlet port side of the auxiliary pump.  
(3) STP Connector (41 pins)  
ꢀꢁ Connected to the STP connection cable.  
(4) Motor Connector (5 pins)  
ꢀꢁ Connected to the motor connection cable.  
(5) Purge Port (KF*2 10)  
ꢀꢁ Introduces a purge gas.  
In order to protect the inside of the STP pump when sucking reactive or  
corrosive gases.  
The STP pump is delivered with a blank flange attached to this port.  
(6) Cooling Water Port (PT*2 (R/C)1/4 Female Screw)  
ꢀꢁ Connected to the STP pump cooling water pipe.  
This port is used when water cooling the STP pump.  
(7) Emergency Vent. Valve  
ꢀꢁ Protects the STP pump.  
It functions immediately after any abnormality/error occurs inside the STP  
pump.  
(8) Ground Terminal  
ꢀꢁ  
Used for grounding.  
Connect the ground cable between this terminal and the ground terminal of the  
STP control unit .  
The ground terminal is marked with label  
.
(9) Temperature Sensor Connector (optional accessory)  
ꢀꢁ A temperature sensor is attached.  
For use with the Temperature Management System (TMS) unit only.  
*1  
*2  
:
:
JVIS  
JIS  
3-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
STP-H600/H1000 Series Instruction Manual  
( ) Inlet Port Flange  
1
( ) Temperature Sensor  
9
Connector  
Optional Accessory  
( ) Outlet Port Flange  
2
( ) Cooling Water Port  
6
( ) Cooling Water Port  
6
( ) STP Connector  
3
( ) Motor Connector  
4
( ) Ground Terminal  
8
( ) Emergency Vent Valve  
7
( ) Purge Port  
5
.
Figure 3.1 Configuration of the STP Pump  
3-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
3.2 Precautions Before Installation  
3.2.1 Operating Environment  
CAUTION  
Chlorine or fluorine system gases can be used in chemical specific pumps (type  
C). When you use gases including alkaline metals, but excluding Li, gases  
including Ga, Hg, Sn, or HBr, contact Seiko Instruments.  
NEVER use corrosive gases (chlorine, fluorine, or other system gases) in the  
STP-H600/STP-H1000 pump or other models without anti-corrosion treatment  
(see Section 1.1, "Usable Gases").  
Install the STP pump in a place meeting the following requirements:  
Ambient Temperature  
Ambient Relative Humidity  
Environment  
0 to 40 °C  
30 to 95 % (no dew condensing)  
ꢀꢁ A place free of externally-applied mechanical shock.  
ꢀꢁ A place free of a heat source  
(Keep clean of the heat source or attach a thermal  
shield plate).  
ꢀꢁ A place free of a strong magnetic field  
(Range: up to 15mT (150G) in the axial direction,  
and up to 3mT (30G) in the radial direction with  
respect to the rotational axis of the STP pump).  
ꢀꢁ A place free of a strong electric field.  
ꢀꢁ A place free of exposure to radiation.  
ꢀꢁ No discharge of high voltage (more than 500 V)  
(If more than 500 V is discharged, contact Seiko  
Instruments).  
STP Pump Installation  
Equipment Conditions  
ꢀꢁ Install the STP pump securely so that foreign  
particles will easily fall into the STP pump (Ex.: Si  
wafers or samples are positioned above the STP  
pump) (To prevent foreign particles from falling into  
the STP pump, design a shield plate with large  
conductance).  
3-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
3.2.2 Installation Area  
Leave enough space for the following in addition to that for the STP pump:  
Sꢀ Space for maintenance and inspection  
Sꢀ Space for connecting cables  
The minimum bending radius of the STP connection cable is 150 mm (see  
Figure 16.1, "External Appearance of the STP Pump" [bending dimensions of  
the STP connection cable]).  
DO NOT excessively bend the cables and beware of any obstacles when  
installing the STP pump.  
Also, leave enough space to install other cables without bending them  
excessively.  
3.2.3 Bench  
A bench must be prepared by the customer to secure the STP pump. The shape  
and size of the bench differ depending upon the type of STP pump. Follow the  
precautions of the WARNING, CAUTION, or NOTICE (See Section 3.3.3, "How to  
Secure the STP Pump").  
The STP pump is provided with a high-speed rotor. Any internal  
abnormality/error may result in a jump in rotational torque leading to personal  
injury or peripheral equipment damage.  
Design and secure the bench for the STP pump so that it can withstand the  
maximum torque generated due to the occurrence of an abnormality/error.  
Refer to Section 3.3.3 “How to Secure the STP Pump” for abnormal torque.  
Secure the customer-prepared bench and the vacuum equipment on the floor or  
peripheral equipment and other equipment in accordance with the customer  
application. NEVER move them while the STP pump is in operation.  
Use fitting bolts with a strength equal to or higher than SUS 304*1 . (Tensile  
strength class*1: 50 or more as a target)  
The screw hole for leg for securing the STP pump is M12*1, and the depth is  
24mm (8 positions).  
For the external appearance of the STP pump, see Figure 16.1, "External  
Appearance of the STP Pump."  
*1 :JIS  
3-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
STP-H600/H1000 Series Instruction Manual  
3.3 How to Install the STP Pump  
Install the STP pump to the vacuum equipment as shown in Figure 3.2.  
The STP pump can be  
installed through the damper  
(optional accessory).  
Vacuum Equipment  
Inlet Port  
Inlet Port Flange  
Vacuum Valve  
Power Cable  
Emergency  
Vent.Valve  
To Power Supply  
STP Pump  
Dry Pump,etc.  
Auxiliary  
Pump  
Emergency  
Vent. Valve Cable  
STP  
Control Unit  
STP Connection Cable  
Motor Connection Cable  
Ground Cable  
Outlet Port  
Leg  
Remote Cable  
To Vacuum Equipment Control Circuit  
Figure 3.2 Installation of the STP Pump to the Vacuum Equipment  
CAUTION  
Chlorine or fluorine system gases can be used in chemical specific pumps (type  
C). When you use gases including alkaline metals, but excluding Li, gases  
including Ga, Hg, In, or Sn, or HBr, contact Seiko Instruments.  
NEVER use corrosive gases (chlorine, fluorine, or other system gases) in the  
STP-H600/H1000 pump or other models without anti-corrosion treatment (see  
Section 1.1, "Usable Gases").  
When you use the STP pump in a place subjected to exposure to radiation,  
contact Seiko Instruments.  
DO NOT open the STP pump through the flange to atmospheric air while the  
STP pump is running.  
If atmospheric air flows into the STP pump, it may not function normally.  
Depending upon the type of the auxiliary pump used, atmospheric air may  
reverse flow into the STP pump when the auxiliary pump stops. Attach a  
vacuum valve to the middle of the piping between the STP pump outlet port  
flange and the auxiliary pump, and close the vacuum valve when the auxiliary  
pump stops.  
3-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
NOTICE  
The STP pump cannot be used with the outlet port open to atmospheric air.  
Always use the auxiliary pump (dry pump or similar one).  
Use an auxiliary pump with a pumping speed of 240 L/min or more.  
Depending upon the type of the auxiliary pump used, oil vapor may contaminate  
the inside of the STP pump. Some oil viscosity could cause a malfunction  
when there is a strong reverse flow of oil.  
Take the following measures to ensure the correct flow of oil:  
Sꢀ Attach a vacuum valve to the middle of the piping between the STP pump  
outlet port flange and the auxiliary pump.  
Sꢀ Attach an absorption trap adjacent to the vacuum valve.  
3.3.1 Cleaning the Seal  
Inspect the seals of inlet and outlet port flanges for dirt or oil spots before installing  
the STP pump in the vacuum equipment.  
Take the following measures for cleaning the seals:  
Sꢀ Clean off with a pure gas.  
Sꢀ Wipe with proper solvent (such as alcohol).  
CAUTION  
A splinter shield is attached to the inlet port flange to prevent foreign particles  
from falling into the STP pump.  
Always leave the splinter shield attached during operation.  
NOTICE  
The splinter shield cannot perfectly prevent foreign materials from falling into the  
STP pump.  
DO NOT install the STP pump in such a manner that foreign materials can  
easily fall into it (for example, Si wafers or samples are positioned above the  
STP pump). If installing the STP pump in such a manner, always attach a  
shield plate with sufficient conductance above the STP pump to prevent foreign  
materials from falling into it. Foreign materials falling into the STP pump  
through the splinter shield may result in product damage.  
Be careful not to scratch the flange of the STP pump.  
Check whether or not there are scratches on the surface, before installing the  
STP pump.  
3-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
3.3.2 STP Pump Installation Positions  
The STP pump can be installed vertically, horizontally, upside-down and slanted.  
Upside-down  
Horizontal  
STP Pump  
Slanted  
Vertical  
Figure 3.3 STP Pump Installation Positions  
When installing the STP pump in a horizontal or slanted position, it is  
recommended to install it so that the direction of the outlet port is on a vertical or  
horizontal plane in the direction of the gravity.  
This makes it possible to reduce the load on the magnetic bearing and the heat  
generated by the STP pump.  
Outlet Port  
Outlet Port  
Direction of Gravity  
Outlet Port  
Outlet Port  
Figure 3.4 Positions of the Outlet Port on the Horizontally or Slanted Installed STP Pump  
3-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
3.3.3 How to Secure the STP Pump  
The STP pump is provided with a high-speed rotor. The worst-case failure may  
result in a jump in rotational torque leading to personal injury or peripheral  
equipment damage.  
The method of securing the STP pump will depend on the installation  
requirements. Secure the STP pump to the vacuum equipment as follows:  
Design and secure the mounting for the STP pump so that it can withstand the  
maximum rotational torque. Refer to Table 3.2 for torque in pump abnormality.  
with bolts  
Refer to 1) When securing the  
inlet port with bolts”  
No  
Secure the inlet port  
Refer to 2) When securing the  
inlet port flange with claw  
clamps”  
Damper  
with Claw Clamps  
Refer to 3) When installing the  
damper in the inlet port flange”  
Yes  
In some cases, the damper and the claw clamper securing cannot be used  
depend on the type of the STP pump.  
Refer to Table 3.1 for tightening torque of the bolt.  
Table 3.1 Tightening torque of bolt  
Size of bolt  
M8  
Tightening torque of bolt (Nm)  
12.0  
24.1  
42.1  
M10  
M12  
When making the leg to secure the base, make them shortened more than ones  
attached to the STP pump.  
Use a material that has a tensile strength of 600N/mm2 or more.  
When securing the base, use stainless steel securing bolts with a tensile  
strength class is 70 or more.  
When using any securing method other than that specified in this manual,  
contact Seiko Instruments.  
3-8  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
STP-H600/H1000 Series Instruction Manual  
1) When securing the inlet port with bolts  
Refer to Table 3.2 for torque in pump abnormality and recommended securing bolts.  
Secure the inlet port flange with all of the boltholes of the size specified in the Inlet  
Port Flange Standard.  
Secure the base with all 8 screw-holes for legs or all 8 attached legs.  
Follow “CAUTION” on page 3-8 about legs and bolts for securing the base.  
Make sure that the recommended securing bolt may be different depending on the  
method of securing the base.  
Table 3.2 Maximum Torque predicted and Recommended securing bolt for inlet port flange  
Model of TMP  
Type of flange  
STP-H600  
ISO160F/ISO160  
2.2×104  
VG150  
2.2×104  
ICF203  
2.2×104  
Torque in pump  
abnormality [Nm]  
Base(8 positions)securing  
No  
Yes  
No  
Yes  
No  
Yes  
Type of bolt  
Standard  
Standard  
Stainless steel  
70 or more  
Standard  
Standard  
Stainless steel  
70 or more  
Standard  
Standard  
Stainless steel  
70 or more  
Recommended  
Carbon steel  
Alloyed steel  
Carbon steel  
Alloyed steel  
Carbon steel  
Alloyed steel  
Type of steel*1  
Strength*1  
securing bolt  
for TMP Flange  
12.9 or more  
12.9 or more  
12.9 or more  
Model of TMP  
STP-H1000  
ISO200F/ISO250F/ISO200  
Type of flange  
VG200  
1.9×104  
ICF253  
2.2×104  
Torque in pump  
abnormality [Nm]  
1.9×104  
Base(8 positions)securing  
No  
Yes  
No  
Yes  
No  
Yes  
Type of bolt  
Standard  
Standard  
Standard  
Standard  
Standard  
Standard  
Recommended  
securing bolt  
Type of steel*1  
Stainless steel Stainless steel Stainless steel Stainless steel Stainless steel Stainless steel  
70 or more 70 or more 70 or more 70 or more 70 or more 70 or more  
for TMP Flange  
Strength*1  
Recommended fitting bolt for  
flange  
Secure the base  
(a) When the base is not secured  
(b) When the base is secured  
Figure 3.5 Example of securing the STP pump  
(When securing the inlet port with bolts)  
*1 Refer to ISO898-1(JISB1051), ISO3506(JISB1054) and AMS6119(Aerospace Material Specification)  
3-9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
STP-H600/H1000 Series Instruction Manual  
2) When securing the inlet port flange with claw clamps  
When securing the inlet port flange with only the claw clamp, the vacuum  
equipment cannot withstand the maximum rotational torque generated by the  
worst-case failure. To make the vacuum equipment withstand abnormal torque,  
secure the base with all 8 screw-holes for legs or all 8 attached legs.  
Follow “CAUTION” on page 3-8 about legs and bolts for securing the base.  
For the claw clamp-type, use the required number of claw clamps as specified in  
Table 3.3. Position the claw clamps evenly on the circumference.  
Table 3.3 Number of Claw Clamps by Size of Flange  
Size of Flange  
ISO 160 or less  
ISO 200 to 250  
ISO 320 or more  
Number of Claw Clamps  
4 or more  
6 or more  
8 or more  
Vacuum Equipment  
Claw Clamps  
Secure the base  
Figure 3.6 Example of securing the STP pump  
(When securing the inlet port flange with claw clamps)  
3-10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
STP-H600/H1000 Series Instruction Manual  
3) When installing the damper in the inlet port flange  
In case of using a damper, secure the base with all 8 screw-holes for legs or all 8  
attached legs. Follow CAUTION” on page 3-8 about legs and bolts for securing  
the base.  
When the base cannot be secured because of the equipment design, install the pump  
with a torque restraint like the one shown in Figure 3.7 (b).  
Use a damper only at the vertically upright position.  
DO NOT remove the bolts and nuts attached to reinforce the damper.  
Vacuum Equipment  
Vacuum Equipment  
Damper Clamps  
Secure the base  
Hole to prevent  
from rotating  
Leg  
(a) When securing the base  
(b) When installing not to rotate  
Figure 3.7 Example of securing the STP pump  
(When installing the damper in the inlet port flange)  
3-11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
STP-H600/H1000 Series Instruction Manual  
3.3.4 Vacuum Piping  
DO NOT open the STP pump through the flange to atmospheric air while the  
STP pump is running.  
If atmospheric air flows into the STP pump, it may not function normally.  
Depending upon the type of the auxiliary pump used, atmospheric air may  
reverse flow into the STP pump when the auxiliary pump stops.  
Attach a vacuum valve to the middle of the piping between the STP pump outlet  
port flange and the auxiliary pump, and close the vacuum valve when the  
auxiliary pump stops.  
In order to let the STP pump bring its performance into full play, follow the  
precautions below:  
1) Be careful not to scratch the flange of the STP pump.  
Before installing the STP pump, check whether or not there are scratches on  
the surface.  
2) Use steel or aluminum tubes with a low gas loss to connect the vacuum  
equipment to the STP pump.  
3) Take measures for minimizing leakage. It is also necessary to degrease the  
tubes as regularly as possible to keep the gas loss as low as possible.  
4) It is recommended to use an auxiliary pump of 240 L/min or more. However,  
the pressure at the inlet and outlet ports varies with the flow rate of gas,  
capacity of the vacuum equipment, length and material of the piping. Select  
an auxiliary pump in accordance with the capacity and starting method  
(simultaneous starting, starting after generating roughing vacuum) suitable for  
the vacuum equipment you use.  
5) Connect the STP pump and the auxiliary pump using stainless steel or  
aluminum alloy tubing, flexible tubing, vacuum rubber or Teflon tubing, etc.  
The following measures can be used to avoid the transmission of the vibration  
of the auxiliary pump to the STP pump and the vacuum equipment.  
Sꢀ DO NOT place the auxiliary pump on the same floor as the vacuum  
equipment.  
Sꢀ Locate the auxiliary pump on a vibration-proof table.  
Sꢀ Attain 1/3 or less of the rotational speed of the auxiliary pump,  
when  
adjusting the inherent frequency of the auxiliary pump installed on a  
vibration-proof table.  
Sꢀ Attach a weight to the piping from the auxiliary pump, or secure the piping  
to a rigid, heavy object free of vibration.  
Sꢀ Use a tube of high flexibility.  
6) Depending upon the type of the auxiliary pump used, oil vapor may  
contaminate the inside of the STP pump. Some oil viscosity could cause a  
malfunction when there is a strong reverse flow of oil.  
Take the following measures to ensure the correct flow of oil:  
Sꢀ Attach a vacuum valve to the middle of the piping between the STP pump  
outlet port flange and the auxiliary pump.  
Sꢀ Attach an absorption trap adjacent to the vacuum valve.  
3-12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
(1) Piping at the Inlet Port Flange  
Attach the inlet port to the high vacuum side.  
Maximum working pressure (Pressure at the inlet port flange applicable  
continuously)  
ꢀꢁ 67 Pa [0.5 Torr] (for water cooled)  
ꢀꢁ 5.3 Pa [4×10-2 Torr] (for air cooled)  
ꢀꢁ 0.9 Pa [7×10-3 Torr] (for natural air cooled)  
ꢀꢁ 13 Pa [0.1 Torr] (when TMS unit is used)  
(2) Piping at the Outlet Port Flange  
Attach the outlet port to the inlet port flange of the auxiliary pump (primary  
side pump).  
Allowable backing pressure (Pressure at the outlet port flange applicable  
continuously)  
ꢀꢁ 400 Pa [3 Torr] (for water cooled, when TMS unit is used)  
ꢀꢁ 267 Pa [2 Torr](for air cooled, natural air cooled)  
To attain the ultimate pressure shown in Table 16.1, "Specifications for STP  
Pump", set the pressure at the outlet port flange to 1.3 Pa (10-2 Torr) or less.  
3-13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
3.3.5 Connecting the Emergency Vent. Valve  
The emergency vent. valve stops the STP pump by introducing gases if any  
abnormality/error occurs in the STP pump.  
Connect the emergency vent. valve (contained in the attached accessories) to the  
purge port as shown in Figure 3.7. Note that the side of the emergency vent. valve  
without filter is connected to the purge port.  
Connect the cable for the emergency vent. valve to the connector, LEAK VALVE  
CON8A, of the STP control unit.  
Always attach the emergency vent. valve.  
DO NOT close the port of the emergency vent. valve (filter side) with a blank  
flange or other type of device.  
The allowable gas pressure ranges from zero [atmospheric pressure] to  
4.9x104Pa [gauge pressure] (zero [atmospheric pressure] to 0.5kgf/cm2 [gauge  
pressure]).  
Use a dry N2 gas or other.  
3.3.6 Connecting the Purge Port  
When sucking reactive or corrosive gases, introduce a dry N2 gas or other gas into  
the STP pump in order to protect the inside of the STP pump.  
As shown in Figure 3.7, introduce a dry N2 gas through the electromagnetic vent.  
valve, needle valve or similar valve (must be prepared by the customer) from the  
purge port.  
For instructions on how to introduce the purge gas, See Section 7.1, "Gas Suction."  
The proper amount of gas purge is approx. 3.4×10-2 Pa·m3/s (20 SCCM).  
The allowable gas pressure ranges from zero [atmospheric pressure] to  
4.9x104Pa [gauge pressure] (zero [atmospheric pressure] to 0.5kgf/cm2 [gauge  
pressure]).  
When not introducing the purge gas, close the purge port with the blank flange  
(attached at delivery).  
3-14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
Dry N2 Gas or Other  
Filter Side  
KF10  
Emergency Vent. Valve  
Dry N2 Gas or Other  
Purge Port  
Figure 3.8 Connecting the Purge Port and Emergency Vent. Valve  
3.3.7 Connecting the Ground Cable  
Connect the ground cable (yellow/green) between the ground terminal of the STP  
pump and the ground terminal of the STP control unit.  
When the resistance between the ground terminals is lower than 0.1 , it is not  
necessary to connect the ground cable after installing the STP pump and the STP  
control unit.  
When the resistance between the ground terminals is over 0.1 , always  
connect the ground cable.  
3-15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
4 Installation of the STP Control Unit  
4.1 Name and Function of Each Part  
4.1.1 Front Panel  
(1) "POWER ON/OFF" Switch (illuminated alternate push button switch, green  
LED)  
ꢀꢁ Press this switch to power ON/OFF the STP pump (MANUAL operation  
only).  
ꢀꢁ The POWER ON/OFF built-in lamp lights when the power is ON.  
(2) "MOTOR START" Switch (momentary push button switch, black)  
ꢀꢁ Press this switch with power ON to rotate and accelerate the STP pump  
(MANUAL operation only).  
ꢀꢁ The "ACCELERATION" lamp lights simultaneously.  
(3) "MOTOR STOP" Switch (momentary push button switch, red)  
ꢀꢁ Press this switch to decelerate and stop the STP pump (MANUAL  
operation only).  
ꢀꢁ The "NORMAL OPERATION" lamp or the "ACCELERATION" lamp goes  
out and the "BRAKE" lamp lights simultaneously.  
When the rotational speed is less than approx. 2000 rpm while the STP  
pump is accelerating, the "BRAKE" lamp does not light.  
(4) "HEATING ON/OFF" Switch (illuminated alternate push button switch, green  
LED)  
ꢀꢁ Press this switch to control the power supplied to the baking (MANUAL  
operation only).  
ꢀꢁ While the power is being supplied to the baking heater, the "HEATING  
ON/OFF" switch built-in lamp lights.  
ꢀꢁ Functions only under the NORMAL OPERATION state (for details, see  
Section 8, "Remote Input/Output Signal Terminal Blocks").  
(5) "NORMAL OPERATION" Lamp (green LED)  
ꢀꢁ Lights during rated operation (NORMAL OPERATION state).  
(6) "ACCELERATION" Lamp (green LED)  
ꢀꢁ Lights during acceleration (ACCELERATION state).  
(7) "BRAKE" Lamp (yellow LED)  
ꢀꢁ Lights during braking (BRAKE state).  
4-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
(8) "OVER TEMPERATURE" Lamp (red LED)  
ꢀꢁ Lights when any of the following abnormalities occurs:  
a) When the motor or electromagnet overheats (110 or higher).  
b) When the STP connection cable is not connected.  
(9) "BATTERY OPERATION" Lamp (red LED)  
ꢀꢁ Lights while the power is being supplied from the battery to the STP pump  
during a power failure.  
(10) "FAILURE" Lamp (red LED)  
ꢀꢁ Lights when any of the following abnormalities occurs:  
a) When the inside of the STP control unit overheats (90 or more  
inside the heat sink).  
b) When the motor or electromagnet overheats (110 or more).  
c) When an abnormality occurs inside the inverter (overload, overspeed).  
d) When the battery is thoroughly worn out and cannot be charged.  
e) When the STP connection cable is not connected.  
(11) "EMERGENCY OPERATION" Lamp (red LED)  
ꢀꢁ Lights when any of the following abnormalities occurs:  
a) When a power failure occurs.  
b) When continuous vibration impact is applied to the rotor causing it to  
come into contact with the touch down bearing.  
c) When the STP connection cable is not connected.  
For details concerning lamps (8) to (11) and abnormalities, see Section 6, "Safety  
Functions When an Abnormality/Error Occurs" and Section 15, "Troubleshooting."  
(12) ROTATION Meter (tachometer)  
ꢀꢁ Indicates the rotational speed (rpm).  
ꢀꢁ The needle moves to the black with an increase in rpm.  
ꢀꢁ The needle moves to the red with a decrease in rpm.  
ꢀꢁ The needle is located in the black during the rated operation.  
4-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
(5)  
(6)  
(12)  
(7)  
(8)  
(9)  
(10)  
(11)  
(4)  
(3)  
(1)  
(2)  
Figure 4.1 STP Control Unit Front Panel  
4-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
4.1.2 Rear Panel  
CAUTION  
A hazardous live voltage may exist at connector/terminal that marked  
DO NOT touch the terminal. Doing so may result in electric shock.  
.
When operating connection/disconnection to terminal, always power OFF the  
STP pump (Switch the breaker "OFF").  
(13) AC POWER Terminal Block (TB1)  
A maximum voltage : Equal to the input voltage of this terminal block. (MAX  
240VAC)  
ꢀꢁ For primary power input.  
(14) MAIN POWER Breaker  
ꢀꢁ Switches ON/OFF the primary power.  
ꢀꢁ A metal fitting is attached to secure the breaker at the OFF position.  
(15) P. CONNECTOR (CON5A)  
A maximum voltage : 60VDC  
ꢀꢁ For connection of the STP connection cable.  
(16) INVERTER OUTPUT Connector (CON2)  
A maximum voltage : 55VDC  
ꢀꢁ For connection of the motor connection cable.  
(17) LEAK VALVE Connector (CON8A)  
A maximum voltage : 60VDC  
ꢀꢁ For connection of the emergency vent. valve cable.  
(18) Ground Terminal  
ꢀꢁ For connection of the ground cable between the STP pump and the STP  
control unit.  
(19) I/O TB2 Terminal Block  
A maximum voltage : Equal to the input voltage of this terminal block. (MAX  
250VAC)  
ꢀꢁ For remote control.  
(20) DC I/O TB3 Terminal Block  
A maximum voltage : 12VDC  
ꢀꢁ For remote control.  
(21) START STOP TB5 Terminal Block  
A maximum voltage : 12VDC  
ꢀꢁ For remote control.  
4-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
(22) START/STOP TB6 Terminal Block  
A maximum voltage : Equal to the input voltage of this terminal block. (MAX  
250VAC)  
ꢀꢁ For remote control.  
For details concerning remote control terminal blocks (19) to (22), see Section 8,  
"Remote Input/Output Signal Terminal Blocks."  
(23) "MANUAL/REMOTE" Changeover Switch  
ꢀꢁ When setting to MANUAL, only start, stop or other operations can be  
performed with the switches on the STP control front panel.  
ꢀꢁ When setting to REMOTE, only start, stop or other operations can be  
performed by inputting the remote signal.  
(24) EXT. BATTERY (CON9) Connector  
A maximum voltage : 55VDC  
ꢀꢁ For connection of an external battery.  
CAUTION  
When using EXT. BATTERY (CON9) connector, always disconnect the internal  
battery connector (see Section 10, "External Battery" for details).  
(21)  
(23)  
(22)  
(24)  
(14)  
(18)  
(16)  
(13) (15)  
(17) (20) (19)  
Figure 4.2 STP Control Unit Rear Panel  
4-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
4.1.3 Inside of the STP Control Unit  
(25) Inverter  
ꢀꢁ This is a three-phase transistor-inverter which starts/stops the STP pump.  
(26) Fuses 3 to 9, 11, and 12  
ꢀꢁ These fuses protect as follows:  
F3 and 4: 250 V, 5 A (for control power protection)  
F5 to 7 : 250 V, 10 A, arc-extinguishing fuses  
(for I/O TB2 terminal block power optional drive protection)  
F8 and 9: 250 V, 10 A, arc-extinguishing fuses  
(for internal and external batteries protection)  
F11 and 12: 250 V, 0.1 A (for START/STOP TB6 terminal block protection)  
(27) Control Circuit Boards  
ꢀꢁ For control of the magnetic bearing, motor, safety functions, etc.  
(28) Internal Battery  
ꢀꢁ For backup during a power failure.  
(29) Internal Battery Connector  
ꢀꢁ For connection of internal battery (when using an internal battery, connect  
it to this connector).  
ꢀꢁ The internal battery has been installed upon shipment of the STP control  
unit so that the customer can use it (the internal battery connector is  
connected to the internal battery).  
CAUTION  
When using an external battery, disconnect internal battery connector, then  
connect external battery connector (see Section 10, "External Battery" for  
details).  
(30) Air Cooling Fan  
ꢀꢁ For cooling the inside of the STP control unit.  
(31) "BATTERY NG" Lamp  
ꢀꢁ Lights when the battery capacity reduces to the minimum level (see  
Section 9.4, "BATTERY NG Lamp).  
(32) "Reset" Switch  
ꢀꢁ Resets the state of inability to start the STP control unit due to a decrease  
in the battery capacity (see Section 9.4, "BATTERY NG Lamp).  
4-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
(25)  
(30) (26) (27) (28) (29)  
(30)  
(32) (31)  
Figure 4.3 Inside of the STP Control Unit  
4-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
4.2 Precautions Before Installation  
4.2.1 Operating Environment  
Install the STP control unit in a place meeting the following requirements:  
Ambient Temperature  
0 to 40 ℃  
Ambient Relative  
Humidity  
30 to 95% (no dew condensing)  
Environment  
ꢀꢁ A place free of exposure to direct sunlight.  
ꢀꢁ A place free of high humidity.  
ꢀꢁ A place free of dust.  
ꢀꢁ A place free of salty air.  
ꢀꢁ A place free of dripping water.  
ꢀꢁ A place free of explosive or inflammable gas.  
ꢀꢁ A place free of corrosive gas.  
ꢀꢁ A place free of radiation.  
ꢀꢁ A place free of strong magnetic and electric fields.  
ꢀꢁ A place free of excessive vibration.  
ꢀꢁ A place free of a source of electric noise.  
ꢀꢁ Install the STP control unit horizontally (within  
M10L).  
Installation Condition  
4-8  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
4.2.2 Installation Area  
Leave enough space for the following in addition to that for the STP control unit.  
Sꢀ Space for maintenance and inspection  
Sꢀ Space for inlet and outlet of air for cooling  
- Top and side:  
- Bottom:  
5 cm or more  
2.4 cm or more (height of the rubber foot)  
Sꢀ Space for connecting the cables  
- Rear: 15 cm or more  
CAUTION  
The minimum bending radius of the STP connection cable is150 mm (see  
Figure 16.2, "External Appearance of the STP Control Unit" [Cable Space]).  
DO NOT excessively bend the cables and beware of any obstacles when  
installing the STP pump.  
Also, leave enough space to install other cables without bending them  
excessively.  
Rack Wall Side  
Rack Wall Side  
Ventilation Port  
Ventilation Port  
Ventilation Port  
5 cm or more 5 cm or more  
5 cm or more 15 cm or more  
Figure 4.4 Peripheral Space of the STP Control Unit  
4-9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
4.3 Attaching the STP Control Unit Front Panel to a Rack  
The dimensions of the STP control unit front panel conform to J IS. Therefore, this panel  
can be attached to any type of commercially-available racks.  
Attach the front panel unit to the rack according to the following steps:  
1) When attaching the front panel to a rack:  
ꢀꢁ Attach the front panel to a rack using the screw holes for the front panel.  
ꢀꢁ Also support the STP control unit from the bottom using a support angle or a similar  
tool.  
2) When attaching the front panel to a movable rack:  
ꢀꢁ Attach the front panel to a movable rack using the screw holes for the front panel.  
ꢀꢁ To protect the STP control unit during transport, remove the rubber foot from the  
bottom and attach the STP control unit to the rack using the screw holes for the  
rubber foot.  
CAUTION  
The STP control unit cannot be supported with only the screws on the front  
panel (the STP control unit is a heavy product).  
Always support it from the bottom. DO NOT block the ventilation port.  
For the peripheral space of the STP control unit, see Figure 4.4, "Peripheral  
Space of the STP Control Unit."  
NOTICE  
For the dimensions of the front panel and positions of screw holes for the rubber  
foot, see Figure 16.2, "External Appearance of the STP Control Unit."  
part: Secure with screw  
Support Angle  
Figure 4.5 Example of Securing the STP Control Unit  
4-10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
4.4 Cable Connection  
4.4.1 Name and Dimensions of Each Cable  
STP Control Unit Side  
STP Pump Side  
Ground Terminal (  
)
Ground Terminal (  
Crimp-Type Terminal Lug (M4*1)  
)
Crimp-Type Terminal Lug (M4*1)  
Yellow / Green Wire  
L 20m : 4mm  
4.3(L 20m)  
φ
6.5(L 20m)  
L 20m : 10mm  
φ
Unit: mm  
Unit: mm  
Unit: mm  
Unit: mm  
24.5  
45 Pin (pin)  
41 Pin (socket)  
5 Pin (pin)  
5 Pin (socket)  
8 Pin (pin)  
Emergency Vent. valve  
Figure 4.6 External Dimensions of Each Cable  
*1  
:
JIS  
4-11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
STP-H600/H1000 Series Instruction Manual  
STP Control Unit Side  
*1  
Crimp-Type Terminal Lug (M4 )  
2.5mm 3core (brown, blue, yellow/green)  
×
(a) without the Power Plug  
Power Side  
STP Control Unit Side  
Plug type  
*1  
Crimp-Type Terminal Lug (M4 )  
NEMA L6-30  
3pin (pin)  
72.3  
Y(N[L2])  
blue  
X(L[L1])  
brown  
2.5mm 3core  
×
G(GND[ ])  
yellow/green  
(brown, blue, yellow/green)  
(b) with the Power Plug*2  
Figure 4.7 External Dimensions of he Power Cable  
*1 : JIS  
*2 : Recognized model to UL3101-1 standard has power cable with the plug.  
4-12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
STP-H600/H1000 Series Instruction Manual  
4.4.2 How to Connect the Cables  
CAUTION  
When connecting/disconnecting cables, always power OFF the STP pump  
(switch the breaker "OFF").  
Failure to do so may result in electric shock or product damage.  
With each connector, align the position of the guide key and insert the pin  
vertically so as not to bend it.  
If the pin is bent, not only may the connector not function normally, but may the  
pin make contact, resulting in a malfunction.  
Lock and securely tighten each connector and screw.  
DO NOT apply voltage to each pin and DO NOT cause any short-circuiting  
between pins.  
Connect each cable securely. DO NOT place heavy objects on the cables nor  
bend them excessively.  
Support each cable so as not to apply direct force to the connectors or terminals.  
If any problem occurs in cables, connectors or terminals, the STP pump may not  
function normally.  
1) Connecting Ground Cable  
Connect the ground cable (yellow/green) between the ground terminal of the  
STP pump and the ground terminal of the STP control unit.  
When the resistance between the ground terminals is lower than 0.1 , it is not  
necessary to connect the ground cable after installing the STP pump and the  
STP control unit.  
CAUTION  
First, connect the ground cable. Next, connect other cables.  
When the resistance between the ground terminals is over 0.1 , always  
connect the ground cable. Failure to do so may result in electric shock.  
4-13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
2) Connecting the STP Connection Cable  
Connect the socket side of the STP connection cable to the STP connector of the  
STP pump and connect the pin side to "P.CONNECTOR CON5A" of the STP  
control unit.  
CAUTION  
Always use the STP pump, STP control unit and the STP connection cables of  
the same model name, serial number and cable length. Failure to do so may  
result in product damage.  
Contact your nearest Seiko Seiki representative if you plan to use units with the  
same model name but different serial numbers and cable length; They must  
be adjusted. In some cases, the configuration may need more adjustment.  
STP-H600/H1000  
Use the STP connection cable that has a label affixed  
Series  
The use of different cables may result in product damage.  
3) Connecting the Motor Connection Cable  
Connect the socket side of the motor connection cable to the motor connector of  
the STP pump and connect the pin side to "INVERTER OUTPUT CON2" of the  
STP control unit.  
CAUTION  
STP-H600/H1000  
Use the motor connection cable that has a label affixed  
Series  
The use of different cables may result in product damage.  
4) Connecting the Emergency Vent. Valve Cable  
Connect the emergency vent. valve to the purge port of the STP pump and the  
connector to "LEAK VALVE CON8A" of the STP control unit (see Section 3.3.5,  
"Connecting the Emergency Vent. Valve").  
4-14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
5) Connecting Primary Power (Primary Side) Cable  
Connect power cable to the "AC POWER TB1" on the STP control unit rear  
panel as shown in Table 4.1.  
Table 4.1 Connecting Primary Power Cable  
TB1 Pin No.  
1 [L(L1)]  
Cable Color  
Brown  
Remarks  
Single phase 200 to 240 V AC M10%  
13A (maximum)  
2 [N(L2)]  
Blue  
Both 50/60 Hz  
Ground  
3 [PE(  
)]  
Yellow/Green  
Secure primary power cable to the STP control unit rear panel using the cable  
fitting tool (see Figure 4.7).  
CAUTION  
Connect the primary power cable securely to prevent incorrect wiring.  
DO NOT apply surge voltage exceeding 1 kV to the input power line.  
Always ground primary power cable to prevent electric shock.  
DO NOT remove the ground cable attached to the terminal block cover.  
4-15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
Magnified  
TB1  
TB1 Cover  
Ground Cable for Cover  
Cable Fitting Tool  
Primary Power Cable  
part: Secure using attached screws  
Figure 4.8 How to Secure Primary Power Cable  
4-16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
5 How to Start/Stop the STP Pump  
CAUTION  
NEVER connect or disconnect any cables while the power is ON.  
NEVER turn the primary power OFF (switch the breaker "OFF") while the STP  
pump is in rotation.  
DO NOT release the inlet port flange or outlet port flange into the atmosphere  
while the STP pump is in rotation.  
5.1 Before Starting  
After completing installation, piping, leakage test of the STP pump and installation of the  
STP control unit, the STP pump is ready for start.  
Check the following items before starting:  
1) Are the STP pump and the STP control unit secured according to the appropriate  
method?  
2) Is the primary power cable connected correctly to the power supply?  
3) Is the power voltage selected properly?  
4) Do the manufacturing No. and the cable length of the STP pump, the STP control unit  
and the STP connection cable match?  
5) Are the labels affixed correctly to the STP connection cable and the motor connection  
cable?  
STP-H600/H1000  
Series  
6) Are the ground cable, the STP connection cable, the motor connection cable, and the  
emergency vent. valve cable securely connected?  
Is the connection cable securely connected?  
Are each terminal and each connector securely locked?  
5.2 Starting/Stopping Time  
Starting time:  
Approx. 6 minutes after start operation.  
Stopping time: Approx. 6 minutes after stop operation.  
If the rotational speed does not attain the rated speed within about 15 minutes, the  
"FAILURE" lamp lights. Before restarting the STP pump, always check whether or not  
there is leakage from the piping or the vacuum equipment and also confirm the capacity and  
the START state of the auxiliary pump.  
NOTICE  
When start/stop operations are frequently performed, the STP pump may  
overheat. Avoid doing so.  
5-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
5.3 Start Procedures  
Start the auxiliary pump before or simultaneously with start of the STP pump.  
Open the vacuum valve located at the outlet port flange side after starting the auxiliary  
pump.  
NOTICE  
DO NOT open the vacuum valve without operating the auxiliary pump.  
Doing so may cause a reverse flow of oil, which could contaminate the inside of  
the STP pump.  
5.4 Stop Procedures  
Close the vacuum valve located at the outlet port flange side just before or after stopping the  
STP pump.  
After closing the valve, stop the auxiliary pump.  
CAUTION  
DO NOT stop the auxiliary pump without closing the vacuum valve. Depending  
upon the type of the auxiliary pump, doing so could cause a reverse flow of  
atmospheric air into the STP pump, which may result in a malfunction.  
NOTICE  
DO NOT stop the auxiliary pump without closing the valve.  
Doing so may cause a reverse flow of oil, which could contaminate the inside of  
the STP pump.  
5-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
There are two methods of operating the STP pump: MANUAL and  
REMOTE.  
Select one which fits your vacuum equipment.  
5.5 Manual Operation  
Slide the "MANUAL/REMOTE" changeover switch on the rear panel to "MANUAL."  
5.5.1 Powering ON  
1) Switch ON the breaker on the rear panel (to prevent incorrect operation, a  
metal fitting is attached to the breaker. Loosen the screw, lift the metal  
fitting and secure it).  
2) Press the "POWER ON/OFF" switch (the switch is thoroughly locked down).  
The magnetic bearing functions and the rotor levitates (POWER ON state).  
The "POWER ON/OFF" switch built-in lamp lights.  
5.5.2 Starting the STP Pump  
1) Press the "MOTOR START" switch on the front panel after powering ON to  
start and accelerate the STP pump.  
The "ACCELERATION" lamp lights (ACCELERATION state).  
2) When the STP pump attains the rated speed of rotations, the  
"ACCELERATION" lamp goes out, and the "NORMAL OPERATION" lamp  
lights (NORMAL OPERATION state).  
5.5.3 Stopping the STP Pump  
1) Press the "MOTOR STOP" switch on the front panel to stop the STP pump.  
The STP pump starts to decelerate and stop.  
The "NORMAL OPERATION" lamp or the "ACCELERATION" lamp goes out,  
and the "BRAKE" lamp lights (BRAKE state)  
(when the rotational speed is less than approx. 2000 rpm while the STP pump  
is accelerating, the "BRAKE" lamp does not light).  
2) The "BRAKE" lamp goes out when the rotational speed slows down to approx.  
1000 rpm.  
5.5.4 Starting the STP Pump after Stopping  
Press the "MOTOR START" switch on the front panel to reaccelerate the STP  
pump.  
5-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
5.5.5 Powering OFF  
1) When all of the "ACCELERATION," "NORMAL OPERATION" and "BRAKE"  
lamps are off:  
Press the "POWER ON/OFF" switch. The switch pops up into the unlocked  
position, the magnetic bearing stops, and the rotor lands (POWER OFF state).  
The "POWER ON/OFF" switch built-in lamp goes out.  
2) When one of the "ACCELERATION," "NORMAL OPERATION" and "BRAKE"  
lamps is lit:  
i. Even when the "POWER ON/OFF" switch is OFF, the rotor will not land.  
Also, operation of the STP pump remains unchanged.  
The "POWER ON/OFF" switch built-in lamp remains lit.  
ii. When the STP pump stops under the above state, the magnetic bearing  
stops, and the rotor automatically lands at the time the "BRAKE" lamp  
goes out.  
NOTICE  
The brake of the STP pump disengages at approx. 1000 rpm and the "BRAKE"  
lamp goes out. The STP pump rotates by inertia even after the "BRAKE" lamp  
has gone out.  
Powering OFF immediately after the "BRAKE" lamp goes out or while any lamp  
of the “ACCELERATION," "NORMAL OPERATION," or "BRAKE" lamp is being  
lit causes the touch down bearing to touch the rotor resulting in a noise.  
Frequent POWER OFF operations will expedite the wear of the touch down  
bearing. It is recommended to power OFF the STP pump after it stops  
thoroughly.  
5-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
5.6 Remote Operation  
Slide the "MANUAL/REMOTE" changeover switch on the STP control unit rear panel to  
"REMOTE."  
5.6.1 Powering ON  
Switch ON the breaker on the STP control unit rear panel,  
(to prevent incorrect operation, a metal fitting is attached to the breaker. Loosen  
the screw, lift the metal fitting and secure it ).  
The magnetic bearing functions and the rotor levitates (POWER ON state).  
The "POWER ON/OFF" switch built-in lamp lights.  
5.6.2 Starting/Stopping the STP Pump  
There are three methods of starting/stopping the STP pump. Use one of them.  
Table 5.1 Starting/Stopping the STP Pump During Remote Operation  
Terminal  
Method  
1
Starting the Pump  
Stopping the Pump  
Block  
TB5  
1) Short the circuit between (3) - Open the circuit  
(2) (STOP IN).  
between (3) - (2) (STOP  
2) Short the circuit between (2) - IN).  
(1) (START IN) for 0.3 seconds  
or more.  
However, when inputting the  
signal simultaneously with  
switching ON the breaker on  
the rear panel, continue to  
short the circuit for 5 seconds  
or more.  
2
3
TB5  
TB6  
Short the circuit between (1) - (3)  
(START/STOP IN).  
Open the circuit  
between (3) - (1)  
(START/STOP IN).  
Input 25 to 250 V AC or 24 to 48 V Open the circuit  
DC between (2) - (1) (START/STOP between (2) - (1)  
IN).  
(START/STOP IN).  
With DC, the (+) side can be  
connected to either (2) or (1).  
5.6.3 Starting the STP Pump After Stopping  
Perform the rotation procedures to reaccelerate the STP pump. See Section 5.6.2,  
"Starting/Stopping the STP Pump."  
The STP pump can be reaccelerated even while the STP pump is stopping.  
5-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
5.6.4 Powering OFF  
When all three ("ACCELERATION", "NORMAL OPERATION" and "BRAKE")  
lamps are off, switch the breaker OFF on the rear panel.  
The magnetic bearing stops, and the rotor lands.  
The POWER ON/OFF switch built-in lamp goes out.  
CAUTION  
NEVER switch OFF the breaker on the rear panel while any of the  
"ACCELERATION," "NORMAL OPERATION" and "BRAKE" lamps is being lit.  
If doing so under the above condition, the STP pump may react as if there was a  
power failure.  
NOTICE  
The brake of the STP pump disengages at approx. 1000 rpm and the "BRAKE"  
lamp goes out. The STP pump rotates by inertia even after the "BRAKE" lamp  
has gone out.  
Powering OFF immediately after the "BRAKE" lamp goes out causes the touch  
down bearing to touch the rotor resulting in a noise.  
Frequent POWER OFF operations may expedite the wear of the touch down  
bearing. It is recommended to power OFF the STP pump after it stops  
thoroughly.  
5-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
6 Safety Functions When an Abnormality/Error Occurs  
The STP pump is provided with safety functions for various abnormalities/errors (See Tables  
6.3 and 6.4, "Safety Functions").  
For troubleshooting, see Section 6.2, "Restarting After Any Safely Function Operates" and  
Section 15, "Troubleshooting."  
6.1 Safety Functions  
6.1.1 Power Failure  
<Operation at a Power Failure>  
When power voltage drops below 170 V due to a power failure, internal battery of  
the STP control unit automatically activate to supply power to the STP pump in  
order to maintain normal function of the magnetic bearing (backup operation  
during a power failure).  
1) The STP control unit detects any power failure of 2 seconds or more and  
automatically switches to the BATTERY OPERATION mode and the STP  
pump decelerates. Gases are introduced from the emergency vent. valve at  
approx. 7000 rpm and the rotor lands on the touch down bearing and stops at  
approx. 2000 rpm.  
In the BATTERY OPERATION mode, the "BATTERY OPERATION" lamp and  
the "EMERGENCY OPERATION" lamp light, and an alarm signal is output  
from I/O TB2 ALARM OUT terminals (10) (9), (10) (11) regardless of  
MANUAL or REMOTE OPERATION mode.  
When the rotor lands on the touch down bearing, the "BATTERY  
OPERATION" lamp and the "EMERGENCY OPERATION" lamp go out, and  
the alarm output is reset.  
2) In case of a power failure of less than 2 seconds, the STP control unit does not  
detect it and the STP pump continues rotating.  
6-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
<Operation after a Power Recovery>  
1) MANUAL Operation  
The STP pump continues decelerating even after a power recovery.  
Press the "START" switch to reaccelerate the STP pump.  
2) REMOTE Operation  
Sꢀ When the START signal is input to the REMOTE input terminal at a  
power recovery to reaccelerate the STP pump.  
Sꢀ When the START signal is not input to the REMOTE input terminal after  
the power recovery, the STP pump continues the BRAKE operation.  
NOTICE  
Establish a sequence so that power can be supplied to the STP control unit  
immediately after a power recovery.  
Table 6.1 shows the states of lamps and the REMOTE output signal at a power  
failure.  
Also, Table 6.2 shows operations of the STP pump after a power recovery.  
Table 6.1 State of Lamps and REMOTE Output Signal at Power Failure  
REMOTE output  
LED lamps  
signal (I/O TB2)  
Rotational speed  
Power failure time  
BATTERY OPERATION”  
EMERGENCY  
OPERATION”  
Lights  
(rpm)  
Power ON ALARM  
signal  
signal  
Less than 2000  
2000 or less  
ON  
ON  
Approx. 2 sec. or more  
Less than approx. 2 sec.  
Goes out  
OFF  
OFF  
Does not detect power failure.  
Table 6.2 Operation of the STP Pump after Power Recovery  
MANUAL operation  
REMOTE operation  
STP pump operation START” REMOTE  
after power recovery signal input after  
power recovery  
STP pump operation  
after power recovery  
Power failure time  
Yes  
No  
Reacceleration  
Deceleration/Stop  
Approx. 2 sec. or more Deceleration/Stop  
Less than approx. 2 sec.  
Continues as before.  
6-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
6.1.2 Abnormal State of Magnetic Bearing  
When the magnetic bearing does not function normally due to a breakage of the  
STP connection cable, disconnection of connectors or any abnormality/error of the  
STP control circuit, the rotor falls on the touch down bearing.  
Simultaneously, the emergency vent. valve is opened to introduce gases and the  
STP pump stops, and the "EMERGENCY OPERATION" lamp lights.  
CAUTION  
When an abnormality/error occurs in the magnetic bearing, check the STP  
pump as well as the STP control unit, and contact Seiko Seiki.  
6.1.3 Excessive Vibration  
When serious vibration or mechanical shock causes the rotor to touch the touch  
down bearing (due to external vibration/impact, intrusion of atmosphere or foreign  
materials into the STP pump or rotor imbalance), the emergency vent. valve is  
opened to introduce gases, the STP pump stops, and the "EMERGENCY  
OPERATION" lamp lights.  
6.1.4 Inverter Overload  
When the STP pump does not attain the rated speed within about 15 minutes after  
starting or when the ACCELERATION state remains unchanged during operation  
for about 15 minutes, the inverter stops and the STP pump continues rotating by  
inertia.  
The "FAILURE" lamp lights.  
6.1.5 Overheating Inside the STP pump  
When the temperature of the motor inside the STP pump exceeds 110 due to an  
abnormal baking temperature or overload operation, the STP pump enters the  
STOP operation.  
The "OVER TEMPERATURE" lamp and the "FAILURE" lamp light.  
6.1.6 Overheating Inside the STP Control Unit  
When the temperature inside the STP control unit (at the heat sink) exceeds 90 ℃  
due to a failure in the air cooling fan, external heat source, etc., the STP pump  
enters the STOP operation.  
The "FAILURE" lamp lights.  
6.1.7 Overspeed  
When the rotational speed of the STP pump exceeds 37,500 rpm due to a failure in  
the inverter, the STP pump enters the STOP operation.  
The "FAILURE" lamp lights.  
6-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
6.1.8 Abnormal Battery Voltage  
When battery capacity became less than operating limit value or there was  
connection mistake, there may be the thing that battery abnormality is detected.  
When battery abnormality was detected, the STP pump does not operate even if the  
STP pump START operation is performed.  
The "FAILURE" lamp lights  
(Both the "EMERGENCY OPERATION" lamp and the "BATTERY NG" lamp inside  
the STP control unit front panel may light).  
When battery voltage became less than DC21V, battery abnormal detection  
operates.  
However, when capacity of battery is less than operating limit value largely, there  
may be the thing that battery abnormal detection does not operate.  
CAUTION  
NEVER disconnect the internal or external battery connection cable while the  
STP pump is under POWER ON state.  
NOTICE  
When the "BATTERY NG" lamp inside the STP pump front panel comes on,  
"FAILURE" and "EMERGENCY OPERATION" lamps also come on.  
When the power is turned on again, the "FAILURE" lamp goes out, while the  
"BATTERY NG" and “EMERGENCY OPERATION" lamps still remain lit.  
6.1.9 Failure of Brake  
If the brake is broken, the emergency vent. valve is opened to force the STP pump to  
stop.  
6-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
6.2 Restarting after Any Safety Function Operates  
1) In case of a power failure:  
Establish a sequence so that power can be supplied to the STP control unit immediately  
after a power recovery.  
2) In case the "FAILURE" lamp lights due to an overload of the inverter  
(the STP pump continues rotating by inertia: [FREE RUN state]):  
<MANUAL OPERATION>  
(1) Press the "MOTOR STOP" switch. The "FAILURE" lamp goes out and the FREE RUN  
state is reset.  
(2) Press the "MOTOR START" switch and press the "MOTOR STOP" switch again. The  
brake functions and the STP pump decelerates.  
(3) After the STP pump stops thoroughly (the "BRAKE" lamp goes out: the needle of the  
tachometer is in the red), turn "OFF" the "POWER ON/OFF" switch, switch the breaker  
"OFF," and remove probable causes of the abnormality/error.  
(4) Restart the STP pump and check if it operates correctly.  
<REMOTE OPERATION>  
(1) Slide the "MANUAL/REMOTE" changeover switch on the rear panel to MANUAL.”  
(2) Press the "MOTOR STOP" switch. The "FAILURE" lamp goes out and the FREE RUN  
state is reset.  
(3) Press the "MOTOR START" switch and press the "MOTOR STOP" switch again. The  
brake functions and the STP pump decelerates.  
(4) After the STP pump stops thoroughly (the "BRAKE" lamp goes out: the needle of the  
tachometer is in the red) , switch the breaker "OFF," and remove probable causes of the  
abnormality/error.  
(5) Restart the STP pump and check if it operates correctly.  
<In case the FREE RUN state is not reset after troubleshooting:>  
(1) Check the tachometer on the front panel.  
(2) If the needle is in the red, it means the STP pump has stopped.  
Power OFF and switch the breaker "OFF."  
(3) If the needle is in the black, the STP pump is rotating.  
Wait for it to enter the red. Then, power OFF and switch the breaker "OFF."  
(4) Restart the STP pump and check if operates correctly.  
CAUTION  
DO NOT power OFF and switch the breaker OFF until the needle is in the red.  
NOTICE  
If the FREE RUN state cannot be reset, it takes about 8 to 10 hours to stop the  
STP pump (the needle is in the red) because it rotates by inertia.  
To stop the STP pump quickly, close the vacuum valve at the outlet port flange  
and introduce gas from the purge port into the STP pump.  
6-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
3) Other Cases  
After the STP pump stops thoroughly, power OFF, switch the breaker "OFF" and  
remove the cause of the abnormality/error. Then, restart the STP pump and check if it  
operates correctly.  
NOTICE  
For probable causes and troubleshooting, See Section 15, "Troubleshooting."  
6-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
6.3 Operation of the Emergency Vent. Valve  
The emergency vent. valve is opened immediately after an abnormality/error in the  
magnetic bearing, excessive vibration or any other abnormality/error is found. Gases (dry  
N2 gases or other) are introduced through the emergency vent. valve into the STP pump.  
Attach the emergency vent. valve taking care of the following NOTICE:  
NOTICE  
It is recommended to introduce dry N2 gases to prevent the STP pump and  
vacuum equipment from getting contaminated.  
The allowable gas pressure ranges from zero [atmospheric pressure] to  
4.9x104Pa [gauge pressure] (zero [atmospheric pressure] to 0.5kgf/cm2 [gauge  
pressure]).  
The rotor comes into contact with the touch down bearing, with the emergency  
vent. valve opened. This may damage the rotor or the touch down bearing.  
To stop the STP pump in a very short time, design an appropriate vacuum  
exhaust system that closes the vacuum valve at the outlet port flange when the  
emergency vent. valve is opened.  
This prevents the rotor and the touch down bearing from being damaged.  
The state of the emergency vent. valve can be checked using "DC I/O TB3 L.  
VALVE N. O. OUT Terminal (9)-(10)."  
6-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
Table 6.3 Safety Functions  
[E.V. used herein is the abbreviation for Emergency Vent.]  
Abnormality  
/Error  
Operation state of safety  
functions  
REMOTE  
output signal  
Probable causes of an  
abnormality/error  
Lamp indication  
State of the STP pump  
Power failure Failure time: 2 seconds or  
more (for details, refer to  
Section 6.1.1, "Power  
"BATTERY  
"ALARM"  
"E.V. valve The STP pump stops.  
OPEN  
signal"  
Levitation continues by battery.  
Power failure.  
Power cable breakage.  
Power cable disconnection  
OPERATION"  
"EMERGENCY  
OPERATION"  
The E.V. valve functions at  
approx. 7000 rpm.  
Failure").  
The rotor lands on the touch down  
bearing at approx. 2000 rpm.  
Abnormal  
magnetic  
bearing  
Rotor descends onto the  
touch down bearing.  
"EMERGENCY  
OPERATION"  
"ALARM"  
"E.V. valve The STP pump stops.  
OPEN  
The E.V. valve is opened.  
STP connection cable breakage.  
STP connection cable  
disconnection.  
signal"  
Abnormal control circuit.  
Excessive  
vibration  
Rotor vibration:  
"EMERGENCY  
OPERATION"  
"ALARM"  
"E.V. valve The STP pump stops.  
OPEN  
The E.V. valve is opened.  
External vibration, and impact.  
Intrusion of atmospheric air.  
Intrusion of foreign materials.  
100 m p-p or more.  
signal"  
Overload of  
inverter  
Current of more than 40A  
flows into the inverter  
momentarily or current of  
more than 11.5A flows into  
the inverter continuously  
for 15 minutes.  
"FAILURE"  
"ALARM"  
The inverter stops.  
The STP pump continues to rotate equipment or piping.  
by inertia.  
Leakage from the vacuum  
Insufficient pressure at inlet  
and outlet port sides.  
Failure of starting of the  
auxiliary pump.  
6-8  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
Abnormality  
/Error  
Operation state of safety  
functions  
REMOTE  
output signal  
Probable causes of an  
abnormality/error  
Lamp indication  
State of the STP pump  
The STP pump stops.  
Overheating Temperature of the STP  
inside the  
STP pump  
"OVER  
TEMPERATURE"  
"FAILURE"  
"ALARM"  
Abnormal baking temperature.  
Insufficient cooling.  
Leakage of vacuum equipment  
or piping.  
pump motor: 110 or  
more  
Continuous repetition of  
startup/shutdown.  
Extraordinarily-high  
temperature.  
Overheating  
inside the  
STP control  
unit  
"FAILURE"  
"FAILURE"  
"ALARM" The STP pump stops.  
Failure in the cooling fan.  
Shielding of cooling air inlet  
and ventilation port.  
Extraordinarily-high  
temperature  
90or more in the STP  
control unit and heat sink.  
Rotational speed of the STP  
pump:37,500 rpm or more.  
Overspeed  
"ALARM" The STP pump stops.  
Failure in the inverter.  
Abnormal  
battery  
terminal  
voltage  
Battery terminal voltage:  
42 V or lower*1  
"FAILURE" *2  
"ALARM" The STP pump does not start.  
Incorrect connection of the  
battery.  
Battery capacity became less  
than operating limit value.  
("EMERGENCY  
OPERATION" or  
"BATTERY NG"  
lamp may light)  
*1 : When capacity of battery is less than operating limit value largely, there may be the thing that battery abnormal detection does not operate.  
*2 : When the power is turned on again, this lamp goes out.  
6-9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
STP-H600/H1000 Series Instruction Manual  
7 Gas Suction, Cooling and Baking the STP Pump  
7.1 Gas Suction  
WARNING  
When sucking gases, they may remain in the STP pump. Introduce a purge  
gas and then exhaust all gasses.  
Residual gases in the STP pump may cause an accident when the STP pump is  
removed.  
Confirm the characteristics of gases to be used, referring to the Material Safety  
Data Sheet (MSDS) you obtain from the gas supplier.  
CAUTION  
Chlorine or fluorine system gases can be used in the chemical specific STP-  
H600C/H1000C (type C). When you use gases including alkaline metals, but  
excluding Li, gases including Ga, Hg, In, or Sn, or HBr, contact Seiko Seiki.  
NEVER use corrosive gases (chlorine, fluorine, or other system gases) in the  
STP-H600/STP-H1000 pump or other models without anti-corrosion treatment.  
Cool the STP pump to within the operating pressure range to prevent the STP  
pump from overheating when sucking gases.  
7.1.1 How to Introduce a Purge Gas  
When sucking reactive or corrosive gases, introduce a purge gas to protect the  
inside of the STP pump.  
1) Connect a needle valve or a similar part to the purge port and introduce a dry  
N2 gas or other gas to perform a gas purge (See Section 3.3.6, "Connecting the  
Purge Port").  
2) The proper amount of the gas purge is approx. 3.4×10 -2 Pa·m3/s (20SCCM).  
The allowable gas pressure ranges from zero [atmospheric pressure] to  
4.9x104Pa [gauge pressure] (zero [atmospheric pressure] to 0.5kgf/cm2 [gauge  
pressure]).  
NOTICE  
When not using the purge port, always mount the blank flange (attached at  
delivery).  
High-pressure at the inlet port may result in a noise. This is no  
abnormality/error.  
7-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
7.2 Cooling the STP Pump  
There are two methods for cooling: Water cooling and Air cooling.  
When sucking gases, cool the STP pump in accordance with the operating pressure range.  
Also, when performing baking, always cool the STP pump.  
Select the best one which fits your vacuum equipment.  
7.2.1 Water Cooling Method  
1) Connect the cooling water pipe to the cooling water port in accordance with  
Figure 7.1.  
2) The female screw PT*1 (RC)1/4 is used in the cooling water port (see Figure 16.1,  
"External Appearance of the STP Pump").  
3) Use connection hose of ID (internal diameter) 8 mm, and secure it to prevent  
water leakage.  
4) Use cooling water under the following conditions:  
Amount of water: 2 L/min  
Temperature:  
Water pressure:  
5 to 25 ℃  
2.9×105Pa (3kgf/cm2) or lower  
NOTICE  
Use clean water as much as possible.  
Cooling water containing foreign materials may corrode or clog the cooling water  
pipe.  
When the cooling system is clogged with foreign materials, clogs may possibly  
be removed by feeding cooling water reversely.  
When the STP pump is overheated due to shortage or suspension of water, the  
protective function detects the overheated condition in the STP pump and stops  
the STP pump.  
As a further safety procedure, attach a flow switch to the cooling water exit so  
that the STP pump stops if abnormal cooling water flow occurs (a flow switch is  
available on the market).  
When the STP pump is not to be used for a long period of time or it is to be  
moved after use, introduce compressed air from one side of the inlet/outlet port  
so that no water will remain inside.  
The joint for water cooling unit is made of brass. To prevent corrosion, connect  
the brass joint.  
*1: JIS  
7-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
STP-H600/H1000 Series Instruction Manual  
7.2.2 Air Cooling Method  
When water cooling is not available, an air cooling unit (optional accessory) can be  
used. Follow the steps below:  
1) Remove the water cooling unit (it is attached with the four M8*1 screws).  
2) Attach the air cooling unit using screw holes for the water cooling unit.  
CAUTION  
Check the rated voltage of the air cooling unit before use.  
Purchase a breaker or fuses for the air cooling unit (remote terminal block TB2  
incorporates fuses).  
When connecting the power for the air cooling unit (over 25 V AC) to the  
remote terminal block, secure the cable using the attached cable fitting tool.  
DO NOT apply excessive force to the cable for the air cooling unit.  
Follow the steps below to run the STP pump associated with the air cooling unit  
(read through Section 8, "Remote Input/Output Signal Terminal Blocks," then  
operate them).  
1) Connect the power cable of the air cooling unit between I/O TB2 FAN OUT  
Terminal (5) - (6)."  
2) Input the power supply of the air cooling unit between I/O TB2 AC. POWER  
IN Terminal (19) - (20)."  
3) The air cooling unit functions under the "ACCELERATION" or "NORMAL  
OPERATION" state.  
4) The air cooling unit also stops when the STP pump stops.  
*1  
:
JIS  
7-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
STP-H600/H1000 Series Instruction Manual  
7.3 Baking the STP Pump  
To attain a lower pressure in a shorter time and reduce the exhaust time, bake the vacuum  
equipment and STP pump.  
WARNING  
The surfaces of the STP pump and its peripheral equipment will become  
extremely hot when performing baking. NEVER touch them with bare hands.  
CAUTION  
When baking the STP pump, always cool it to prevent overheating (for the  
cooling method, see Section 7.2, "Cooling the STP Pump").  
Start baking after cooling is started.  
Set the temperature of the baking heater to 120 or lower (an optional baking  
heater is set to 110 or lower).  
DO NOT suck gases during baking to prevent overheating.  
NOTICE  
To exhaust the gas discharged from the vacuum equipment and the inner wall of  
the STP pump, run the STP pump during baking.  
7.3.1 Attaching a Baking Heater  
1) Attach a baking heater (optional accessory) as near as possible to the inlet port  
flange.  
2) Affix the "Hot Surface Warning Label" to the surface of the STP pump so that  
the operator can see it clearly at any time.  
CAUTION  
Check the rated voltage of the baking heater before use.  
Wind the baking heater around the surface of the STP pump tightly.  
If the baking heater is not wound tightly, the loose parts will overheat.  
Procure protective parts for the baking heater, such as a residual current  
operated circuit breaker and fuses when using the baking heater (remote  
terminal block TB2 incorporates fuses).  
DO NOT apply excessive force to the cable for the baking heater.  
7-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
・警告  
WARNING  
Hot surface  
.
Can burn hands  
Do not touch  
高温部  
.
やけどします。  
触らないでく  
ださい。  
.
This label is attached  
to the baking heater  
.
Baking Heater  
(Optional Accessory)  
Cooling Water Port  
Cooling Water Port  
Water Cooling Unit  
(Standard Accessory):  
Used when water cooled  
Air Cooling Unit  
(Optional Accessory):  
Used when air cooled  
Table 7.1 Attaching Positions of the Cooling Unit and Baking Heater  
7-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
7.3.2 Interlocking the STP Pump Operation with Baking  
Follow the steps below to interlock the STP pump operation with baking (read  
through Section 8, " Remote Input/Output Signal Terminal Blocks," then start  
operation).  
CAUTION  
Check the rated voltage of the baking heater before use.  
Always turn OFF the power to the remote terminal block from equipment, before  
connecting/disconnecting the remote wire to the remote terminal block.  
Failure to do so may result in electric shock.  
When connecting the power for the baking heater (over 25 V AC) to the remote  
terminal block, secure the cable using the attached cable fitting tool.  
DO NOT apply excessive force to the cable for the baking heater.  
1) MANUAL OPERATION (See Section 8.1, "I/O TB2 Terminal Block.")  
(1) Connect the power cable of the baking heater between "I/O TB2 HEATER  
OUT Terminals (1) - (2)."  
(2) Input the power supply of the baking heater between "I/O TB2 AC.  
POWER IN Terminals (19) - (20)."  
(3) Press the HEATING ON/OFF switch (an alternate type) under the  
NORMAL OPERATION state (the HEATING ON/OFF switch built-in  
green lamp lights. The heater is "ON" while the lamp is being lit).  
(4) Press the HEATING ON/OFF switch under any state other than the  
NORMAL OPERATION. The HEATING ON/OFF switch built-in green  
lamp lights and the baking heater functions under the NORMAL  
OPERATION state.  
2) REMOTE OPERATION (See Sections 8.1 and 8.2, "I/O TB2 Terminal Block"  
and "DC I/O TB3 Terminal Block.")  
(1) Perform the same steps (i, ii) as for the MANUAL OPERATION.  
(2) Short the circuit between "I/O TB3 HEATING Terminals (2) – (4)" (the  
HEATING ON/OFF switch built-in green lamp lights. The heater is "ON"  
while the lamp is being lit).  
(3) Short the circuit between "I/O TB3 HEATING Terminals (2) – (4)" under  
any state other than the NORMAL OPERATION state.  
(4) The HEATING ON/OFF switch built-in green lamp lights and the baking  
heater functions under the NORMAL OPERATION state.  
NOTICE  
The baking heater functions only while the STP pump is under the NORMAL  
OPERATION state.  
7-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
8 Remote Input/Output Signal Terminal Blocks  
8.1 I/O TB2 Terminal Block  
This is a terminal block for a remote output signal. Use it in accordance with Table 8.1 and  
Figure 8.1. The terminal block functions in both the MANUAL and REMOTE operations.  
The screw for the terminal is M3*1  
Five abbreviations are used in the following tables:  
N.O :Normal Open N.C :Normal Close  
COM :Common  
IN  
L1  
:Input Terminal OUT : Output Terminal  
:Hot N1: Neutral (or Grounded Supply)  
CAUTION  
Always turn OFF the power to the remote terminal block from equipment, before  
connecting/disconnecting the remote wire to the remote terminal block.  
Failure to do so may result in electric shock.  
When connecting optional power (over 25 V AC or over 60 V DC) to terminals  
(1) to (6), (19) and (20), secure the remote cable using the attached cable fitting  
tool (see Figure 8.3).  
DO NOT apply excessive force to the remote cable.  
Table 8.1 I/O TB2 Terminal Block  
Terminal  
Description  
Terminal for inputting power for driving the baking heater,  
auxiliary pump and air cooling unit. The voltage input to  
this terminal is output to the specified terminal block through  
operation of the STP pump.  
(20) AC.POWER (19)  
Connect hot to terminal (19)[L1].  
Connect neutral (or grounded supply) to terminal (20)[N2].  
N1  
IN  
L1  
Terminal for connecting the baking heater.  
In either case of the following two types of operation under  
the NORMAL OPERATION state, voltage input to the AC  
power terminal ((19)-(20)) is output to this terminal.  
1) Turn ON the HEATING switch on the front panel  
through MANUAL Operation.  
(2)  
HEATER  
N.O OUT  
(1)  
2) Short the circuit between I/O TB3 HEATING terminal  
(2)-(4) through REMOTE Operation (See Section 7.3.2,  
"Interlocking the STP Pump Operation with Baking").  
Terminal for starting the auxiliary pump.  
The voltage input to the AC. POWER terminal ((19)-(20)) is  
output to this terminal through the STP pump rotation. DO  
NOT connect the auxiliary pump directly.  
Use separate power and a relay.  
Use another power and a relay. See Figure 8.2.  
(4)  
R.PUMP  
N.O OUT  
(3)  
If the I/O TB3 ((1)-(3)) is opened, it is  
diagnosed as abnormal auxiliary  
pump, the "FAILURE" lamp lights and  
the power is cut OFF.  
NOTICE  
*1  
:
JIS  
8-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
STP-H600/H1000 Series Instruction Manual  
Terminal  
FAN  
Description  
Terminal for connecting the air cooling unit.  
The voltage input to AC. POWER terminal ((19)-(20)) is  
output to this terminal when the STP pump is in acceleration  
or rated operation.  
The voltage is cut OFF when the STP pump is under the  
BRAKE state (See Section 7.2.2, Air Cooling Method").  
(6)  
(8)  
(5)  
(7)  
N.O OUT  
Terminal for outputting the POWER ON state signal.  
This terminal is closed when the magnetic bearing functions  
and the rotor levitates.  
POWER  
N.O OUT  
Terminal for outputting the ALARM signal.  
When an abnormality/error is detected under the POWER ON  
state, the terminal between (10)-(9) is closed, and the  
terminal between (10)-(11) is opened.  
(10) ALARM OUT (9)  
COM  
(12)  
N.O  
N.C  
(11)  
Terminal for outputting the BRAKE state signal.  
This terminal is closed when the STP pump is in brake.  
(14)  
Terminal for outputting the NORMAL OPERATION state  
signal.  
When the STP pump is in rated operation, the terminal  
between (16)-(13) is closed, and the terminal between (16)-  
(15) is opened.  
NORM OUT (13)  
N.O  
COM N.C  
(15)  
(16)  
(18)  
Terminal for outputting the ACCELERATION state signal.  
This terminal is closed when the STP pump is in acceleration.  
POWER  
N.O OUT  
(17)  
8-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
N.O  
N.C  
:Normal Open  
:Normal Close  
:Common  
I/O TB2 :  
Remote Output Signal Terminal Block  
COM  
19  
L1  
AC POWER  
N1  
Power input terminals for  
baking heater, auxiliary pump  
and air cooling unit  
Fuse F5 250 V AC 10A  
Fuse F7  
250 V AC 10A  
20  
1
HEATER  
Terminal for connecting the  
baking heater  
CR1  
Fuse F6  
250 V AC 10A  
N O OUT  
.
2
3
R PUMP  
.
Terminal for the auxiliary pump  
Terminal for the air cooling unit  
CR2  
CR3  
CR4  
N O OUT  
4
5
.
FAN  
N O OUT  
.
6
7
POWER  
POWER ON state  
signal output terminal  
N O OUT  
.
8
9
NO OUT  
.
CR5  
CR5  
ALARM state  
signal output terminal  
ALARM OUT COM  
10  
NC OUT  
.
11  
12  
BRAKE  
NO OUT  
BRAKE state  
signal output terminal  
CR10  
.
14  
13  
NO OUT  
.
CR6  
CR6  
NORMAL OPERATION state  
signal output terminal  
NORM OUT COM  
16  
NC OUT  
.
15  
17  
ACC  
.
ACCELERATION state  
signal output terminal  
CR7  
NO OUT  
.
18  
Figure 8.1 I/O TB2 Remote Output Signal Terminal Block  
8-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
Table 8.2 shows rated contacts for relays CR1 to 4, CR7, and CR10 (See Figure 8.1) and CR8  
(See Figure 8.4).  
Table 8.2 Rated Contacts for Relays CR1 to 4, CR7, CR8 and CR10  
Induction Load  
Resistance Load  
(cos φ=0.4)  
(cos φ=1)  
(L/R=7 ms)  
250 V AC 10 A  
30 V DC 10 A  
250 V AC 5A  
30 V DC 5A  
Rated Load  
Rated Current  
Maximum Contact Point  
Voltage  
10 A  
380 V AC  
125 V DC  
Maximum Contact Point  
Current  
10 A  
Maximum Open/Close  
Capacity  
AC: 2500 VA  
DC: 300 W  
AC: 1250 VA  
DC: 220 W  
Minimum  
Applicable Load  
5 V DC 10 mA  
Table 8.3 shows rated contacts for relays CR5 and 6 (see Figure 8.1).  
Table 8.3 Rated Contacts for Relays CR5 and 6  
Induction Load  
(cos φ=0.4)  
(L/R=7 ms)  
Resistance Load  
(cos φ=1)  
250 V AC 10 A  
30 V DC 10 A  
250 V AC 7.5 A  
30 V DC 5A  
Rated Load  
Rated Current  
Maximum Contact Point  
Voltage  
10 A  
380 V AC  
125 V DC  
Maximum Contact Point  
Current  
10 A  
Maximum Open/Close  
Capacity  
AC: 2500 VA  
DC: 300 W  
AC: 1875 VA  
DC: 150 W  
Minimum  
Applicable Load  
5 V DC 100 mA  
8-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
AC Power  
Relay  
3
4
R.PUMP  
NO.OUT  
R.P  
R.P : Auxiliary Pump  
Figure 8.2 Example of Connecting I/O TB2 R. Pump Terminal  
CAUTION  
When connecting optional power (over 25 V AC or over 60 V DC) to terminals  
(1) to (6), (19) and (20), secure the remote cable using the attached cable fitting  
tool (see Figure 8.3).  
DO NOT apply excessive force to the remote cable.  
DO NOT disconnect the ground cable for the metallic terminal block cover.  
Cable Fitting Tool  
Magnified  
Metallic Terminal Block Cover  
Ground Cable for Cover  
Remote Cable  
Figure 8.3 Example of Securing the Remote Cable  
8-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
8.2 DC I/O TB3 Terminal Block  
This is a terminal block for a remote input/output signal. Use it in accordance with Table  
8.4 and Figure 8.4. Terminals excluding the HEATING terminal function in either case of  
the REMOTE/MANUAL operation.  
The screw for the terminal is M3*1.  
Four abbreviations are used in the following tables:  
N.O :Normal Open  
N.C :Normal Close  
IN  
:Input Terminal OUT :Output Terminal.  
Table 8.4 DC I/O TB3 Terminal Block  
Terminal  
Description  
Terminal for connecting the thermal switch to protect the  
auxiliary pump.  
When the normal close type thermal switch for protecting  
the auxiliary pump is connected between this terminal,  
short the circuit between I/O TB2 R.PUMP terminal (3)-(4)  
and stop the auxiliary pump.  
The "FAILURE" lamp lights.  
When this function is not used, short the circuit between (3)-  
(4) (shorted at delivery).  
(1)  
(3)  
Terminal for baking heater remote control.  
When the circuit between this terminal is shorted through  
the "REMOTE" operation, the voltage for the baking heater  
is output between I/O TB2 HEATER terminal (1)-(2) during  
rated operation.  
(2)  
(4)  
Terminal for inputting the rotation INHIBIT signal. When  
the terminal is set to open, the STP pump does not rotate  
despite the STP pump operation.  
And only the auxiliary pump start voltage is output to I/O  
TB2 R.PUMP terminal ((3)-(4)).  
(5)  
(7)  
When this function is not used, short the circuit between (5)-  
(7) (shorted at delivery).  
Terminal for the emergency vent. valve operation signal.  
This terminal is closed while the emergency vent. valve is  
opened.  
(10) L.VALVE  
N.O OUT  
(9)  
Ground  
(6)  
(8)  
Not-used  
*1  
:
JIS  
8-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
STP-H600/H1000 Series Instruction Manual  
N.O : Normal Open  
N.C : Normal Close  
IN : Input Terminal  
OUT : Output Terminal  
DC I/O TB3 :  
Remote Input Signal Terminal Block  
+12V  
1 k  
1 k  
1 k  
Ω
Ω
Ω
R.P.THER  
N.C IN  
1
3
Auxiliary pump  
thermal switch  
input terminal  
5
7
INHIBIT  
N.C IN  
Rotation inhibit signal  
input terminal  
HEATING  
N.O IN  
2
4
Baking heater control  
signal input terminal  
9
L..VALVE  
N.O OUT  
Emergency vent. valve  
open signal output terminal  
CR8  
10  
6
8
Ground terminal  
Unused terminal  
NOTICE  
Short the circuit between (1)-(3), when the R.P.THER terminal is not used.  
Short the circuit between (5)-(7), when the INHIBIT terminal is not used.  
See Table 8.2 for the contact ratings of relay CR8.  
Figure 8.4 DC I/O TB3 Remote Input Signal Terminal Block  
8-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
8.3 START STOP TB5 Terminal Block  
This is a terminal block for a contact type remote input signal. Use it in accordance with  
Table 8.5 and Figure 8.5 (See Section 5.6.2, "Starting/Stopping the STP Pump").  
The terminal block operates only during the REMOTE operation.  
The screw for the terminal is M4*1  
Three abbreviations are used in the following tables:  
N.O :Normal Open N.C :Normal Close  
IN :Input Terminal  
.
Table 8.5 START STOP TB5 Terminal Block  
Terminal  
Description  
Terminal for the START signal input terminal. The  
following two methods are available:  
1) Short the circuit between terminals ((2)-(3)) for 0.3  
(1)  
(2)  
seconds or more. When this signal is input  
simultaneously with switching ON the breaker on the  
rear panel, short the circuit for 5 seconds or more.  
2) Short the circuit between terminals ((1)-(3)).  
In this case, terminal (2) is not used.  
N.O  
START IN  
Terminal for the STOP signal input terminal.  
The STOP signal has priority for the START signal.  
During the START operation input from terminal  
(2)  
(3)  
(1)-(2) of the START/STOP TB6 terminal block, the STP  
pump does not stop.  
N.C  
1)  
2)  
In case of above 1):  
Rotation will stop when the circuit between this  
terminal is opened.  
STOP IN  
In case of above 2):  
Rotation will stop when the circuit between terminal  
(1)-(3) is opened.  
START STOP TB5 Remote Input Signal Terminal Block  
N.O :Normal Open  
N.C :Normal Close  
IN :Input Terminal  
Contact Type START/STOP Input Terminal  
+12V  
CR9  
In case of 2) in Table 8.5  
In case of 1) in Table 8.5  
1
2
3
START Signal  
N.O START IN  
Input Terminal  
STOP Signal  
Input Terminal  
N.C STOP IN  
Figure 8.5 START STOP TB5 Remote Input Signal Terminal Block  
*1  
:
JIS  
8-8  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
STP-H600/H1000 Series Instruction Manual  
8.4 START/STOP TB6 Terminal Block  
This is a terminal block for a voltage type remote input signal.  
Use it in accordance with Table 8.6 and Figure 8.6 (See Section 5.6.2, Starting/Stopping the  
STP Pump").  
The terminal block operates only during REMOTE operation.  
The screw for the terminal is M4*1  
CAUTION  
When connecting optional power (over 25 V AC) to terminals (1) or (2), secure  
the remote cable using the attached cable fitting tool (See Figure 8.3).  
DO NOT apply excessive force to the remote cable.  
DO NOT disconnect the ground cable for the terminal block cover.  
Table 8.6 START/STOP TB6 Terminal Block  
Terminal  
Description  
Terminal for the voltage type START/STOP signal terminal.  
When the voltage of 25 to 250 V AC or 24 to 48 V DC is  
applied between these terminals, the STP pump starts.  
When the circuit between these terminal is opened, the STP  
pump stops.  
(1)  
(2)  
In case of DC input, the STP pump can start to rotate by  
connecting the plus (+) side either to (1) or (2).  
START/STOP IN  
START/STOP TB6 Remote Input Signal Terminal Block  
Voltage Type START/STOP Input Terminal  
0.1  
120  
Ω
μF  
Equivalent Resistance  
40 k  
Ω
Fuse  
1
2
F11 250V AC 0.1A  
START/STOP IN  
START/STOP Signal 25 - 250 V AC  
Input Terminal 24 - 48 V DC  
F12 250V AC 0.1A  
Figure 8.6 START/STOP TB6 Remote Input Signal Terminal Block  
*1  
:
JIS  
8-9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
STP-H600/H1000 Series Instruction Manual  
9 Internal Battery  
Four batteries (PORTALAC Series Small-Sized Sealed Lead-Acid Batteries, PE12V 2WS :  
12V, 2Ah, manufactured by J apan Storage Battery Co., Ltd.) (hereinafter referred to as the  
internal battery) are inserted into the STP pump for backup during a power failure. These  
are rechargeable. During a power failure, power is supplied from the batteries, and  
levitation by a magnetic bearing continues for 6 minutes while the pump is decelerating and  
the rotor is landing.  
Fully-charged batteries are capable of consecutively backing up the operation of the STP  
pump twice.  
CAUTION  
Do not use unspecified battery.  
If you use unspecified battery, contact Seiko Seiki before use.  
9.1 Life of the Internal Battery  
The life of the internal battery is affected by ambient temperatures.  
Pay attention to ambient temperatures (the operating ambient temperatures range from 0 to  
40 LC).  
Figure 9.1 shows the relationship between the ambient temperatures and the battery life.  
CAUTION  
Replace the battery approx. once a year to make sure it is always ready to back  
up the operation of the STP pump.  
For instructions on how to replace the battery, see Section 9.5, "How to Replace  
the Internal Battery."  
9.2 Allowable Shelf Life of the Internal Battery  
When the internal battery is left uncharged, its capacity will be reduced by self-discharge.  
The allowable shell life of the internal battery is affected by ambient temperatures. Pay  
attention to its storage temperatures (the storage temperatures range from - 20 to + 40 LC).  
Figure 9.2 shows the relationship between storage temperatures and the allowable shelf life  
of the internal battery.  
CAUTION  
When the STP pump is left unused over the specified period, charge the internal  
battery for 90 minutes or more, or replace it with a new one before starting the  
STP pump. For instructions on how to recharge the internal battery, see  
Section 9.3, "How to Charge the Internal Battery." For instructions on how to  
replace the internal battery, See Section 9.5, "How to Replace the Internal  
Battery."  
9-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
50  
40  
10  
30  
20  
5
4
3
2
10  
5
1
0.5  
10  
10  
20  
30 40  
Ambient Temperatures(  
20  
30  
40  
Ambient Temperatures(  
)
)
Figure 9.1 Life of the Internal Battery  
Figure 9.2 Allowable Shelf Life of the Internal  
Battery  
9.3 How to Charge the Internal Battery  
The battery is charged while the power is "ON," regardless of the operation state of the STP  
pump.  
After a power failure, charge the battery for approx. 12 times the BATTERY BACKUP time.  
For example, when the BATTERY BACKUP time is 6 minutes, it takes approx. 75 minutes  
to charge the battery.  
If a power failure occurs 2 times within 90 minutes, charge the battery for more than 90  
minutes before starting to operate the STP pump.  
9-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
9.4 BATTERY NG Lamp  
If battery capacity became less than operating limit value or there was connection mistake,  
the "BATTERY NG" lamp on the rear side of the STP control unit front panel (See 9.3, "Life  
of the Internal Battery)", the "FAILURE" lamp, and the "EMERGENCY OPERATION" lamp  
on the STP control unit front panel light, and the STP pump does not start.  
Replace the battery with a new one and reset the "Battery Capacity Low Warning"  
(BATTERY NG) lamp.  
However, when capacity of battery is less than operating limit value largely, there may be  
the thing that battery abnormal detection does not operate.  
For instructions on how to replace the battery and restart the STP pump, see 9.5 How to  
Replace the Internal Battery "  
CAUTION  
When the " BATTERY NG" lamp lights, always replace the internal battery with  
a new one.  
"BATTERY NG" lamp does not indicate the battery exchange time beforehand.  
There is the case that battery deteriorates even if this lamp does not turn on.  
According to 9.1 "Life of the internal Battery", change battery regularly.  
NOTICE  
When the power is turned on again, the "FAILURE" lamp goes out.  
9-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
9.5 How to Replace the Internal Battery  
WARNING  
When covering the battery case, be sure not to catch the battery cable in the  
battery case. Failure to do so may shortcircuit the battery and result in a  
leakage, production of fire or explosion.  
CAUTION  
Turn OFF the primary power (Switch the breaker "OFF"), before replacing the  
battery.  
Failure to do so may result in product damage or a problem due to electric  
shock or short-circuiting of the internal circuit.  
DO NOT touch any portions other than those designated.  
Careless touch may cause electric shock and/or a short-circuiting of the internal  
circuit, resulting in product damage or a problem.  
DO NOT use a new battery and a worn-out battery simultaneously.  
DO NOT use different types of batteries simultaneously.  
NOTICE  
When replacing the battery, record the next replacement date of the battery on  
the "Battery Instruction Label" (See Figure 9.3) on the front panel.  
The first replacement date of the battery has been specified on the "Battery  
Instruction Label" (See Figure 9.3) on the front panel at delivery.  
9.5.1 How to Replace the Battery Case  
1) Stop the STP pump, and turn OFF the primary power (switch the breaker  
"OFF").  
2) Unscrew the 4 screws attached to front panel (1) and bring the front panel  
down towards you.  
3) Disconnect internal battery connector (2) (CON12).  
4) Unscrew battery case screws (3) and pull out battery case (4) from the STP  
control unit.  
5) Replace battery case (4) containing a new battery.  
6) Tighten battery case screw (3) and secure the battery case (4).  
7) Connect internal battery connector (2) (CON12).  
8) Turn ON the primary power (switch the breaker "ON") as well as the POWER  
switch. Then, check whether or not "BATTERY NG" lamp (7) on the rear side  
of the front panel lights.  
9) If "BATTERY NG" lamp (7) located on the rear of the front panel lights, press  
RESET switch (8) on the side of "BATTERY NG" lamp (7) to reset it (although  
an impact noise may generate momentarily during reset, it is not indicative of a  
problem).  
10) Turn OFF the POWER switch and the primary power (Switch the breaker  
"OFF").  
11) Close front panel (1) and tighten the 4 screws that secure the front panel.  
12) Record the next replacement date (after one year) of the battery on "Battery  
Instruction Label" (9) attached to the front panel (See Figure 9.3).  
9-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
9.5.2 How to Replace the Battery in the Battery Case  
1) Remove front plate (5) from battery case (4) by unscrewing 4 M2 screws.  
2) Disconnect battery internal connector (6).  
3) Remove inside plate(10) from battery case by unscrewing 4 M2 screws and  
pull out the battery from the case (it can be pulled out toward you).  
4) Replace the four worn-out batteries with new ones, the side, from which the  
cord projects, up.  
5) Attach inside plate(10).  
6) Connect battery internal connector (6) and attach front plate (5) to the battery  
case (Be careful not to catch the cable).  
WARNING  
When covering the battery case, be sure not to catch the battery cable in the  
battery case. Failure to do so may shortcircuit the battery and result in a  
leakage, production of fire or explosion.  
NOTICE  
New batteries and battery cases are supplied by Seiko Seiki.  
9-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
(1) STP Control Unit Front Panel  
(2) Internal Battery Connector  
(3) Battery Case Screw  
(4) Battery Case  
(5) Battery Case Front Plate  
(6) Battery Internal Connector  
(7) BATTERY NG Lamp  
(8) RESET Switch  
(9) Battery Instruction Label  
(10) Battery Case Inside Plate  
(11) Flat Head Screw (M2)  
(2)  
(8)  
(7)  
(1)  
(4)  
(10)  
(3)  
(5)  
(11)  
(9)  
CAUTION  
Change batteries every year  
.
・注意  
◇バッテリーは1年毎に交換してく  
ださい。  
Do not use external batteries  
in combination with internal  
◇内部バッテリーと外部バッテリー  
を同時に使用しないでください。  
batteries  
.
Next exchange date  
次回交換日/  
(6)  
Affix this label to the front panel.  
(5)  
Figure 9.1 How to Replace the Internal Battery  
9-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
9.6 How to Dispose of the Internal Battery  
Dispose of or collect worn-out batteries in accordance with the guidelines given by each  
national and/or local government. If you procure batteries, dispose of them in accordance  
with the guidelines given by the battery manufacturer.  
CAUTION  
Pay attention to short-circuiting or leakage of the batteries when disposing of  
them.  
To prevent the batteries from short-circuiting, remove the lead wires and  
connectors from the batteries.  
Cut the lead wires one by one so as not to cause a short-circuiting.  
Coat the conductive part of the batteries with tape.  
NOTICE  
It is recommended to dispose of worn-out batteries after discharging them  
thoroughly.  
Batteries manufactured by Japan Storage Battery Co., Ltd. will be collected by  
Japan Storage Battery Co., Ltd. For detailed information, contact Japan  
Storage Battery Co., Ltd. or Seiko Seiki.  
9-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
10 External Battery  
CAUTION  
DO NOT use (an) external batteries when internal batteries are inserted.  
Failure to do so may result in product damage or a problem.  
It could cause a problem, such as product damage.  
DO NOT connect batteries in parallel.  
Replace batteries with new ones before the end of their life.  
When the STP pump is left unused over the specified period, charge battery for  
90 minutes, or replace it with a new one before starting the STP pump.  
NOTICE  
The life, allowable shelf life, and operation frequency of the battery differ  
depending upon the type of the battery and ambient temperatures. See  
technical data and information prepared by the battery manufacturer.  
10.1 Specifications for the External Battery  
When you use external batteries, select one in accordance with the specifications given in  
Table 10.1.  
CAUTION  
If you use external batteries which do not conform to the specifications, battery  
backup will not function during a power failure which could result in damage to  
the STP pump.  
Table 10.1 Specifications for the External Battery  
Item  
Capacity  
Specifications  
1.8 Ah or more  
48 V  
Voltage  
Current  
2.3 A  
Operation time  
Operation frequency  
Charging voltage  
Approx. 6 minutes  
Continues 2 times or more  
55 V  
It is recommended to use PORTALAC Series Small-Size Sealed Lead-Acid Battery  
manufactured by J apan Storage Battery Co., Ltd. as an external battery.  
10-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
10.2 Installation of the External Battery  
CAUTION  
Turn OFF the primary power (Switch the breaker "OFF"), before connecting the  
battery.  
Failure to do so may result in product damage or a problem due to electric  
shock or short-circuiting of the internal circuit.  
After disconnecting the internal battery connector (CON12), connect the external  
battery connector (CON9).  
Having both connected simultaneously may result in damage to the STP control  
unit.  
DO NOT use new batteries and worn-out batteries simultaneously.  
DO NOT use different types of batteries simultaneously.  
If the battery is not securely connected, backup during a power failure does not  
function.  
Install the external battery in accordance with the following steps:  
1) Stop the STP pump, and turn OFF the primary power (Switch the breaker "OFF").  
2) Unscrew the 4 screws attached to the front panel and bring it down towards you.  
3) Pull out the battery according to the method described in Section 9.5.1, "How to Replace  
the Battery Case," Steps 3) and 4).  
Secure the internal battery connector (CON12) so as not to interfere with other  
parts inside the STP control unit.  
4) Close the front panel of the STP control unit and tighten the 4 screws that secure the  
front panel.  
5) Connect the external battery to the attached external battery plug, paying attention so  
that "+" and "" are positioned correctly. "+" and "" of the external battery plug are  
represented by (1) and (2), respectively.  
CON 9 PIN No. Polarity of Battery  
1
2
+
-
6) Connect the external battery plug to the external battery connector (EXT. BATTERY  
CON9) on the rear panel.  
7) Record the next replacement date of the battery in the "Battery Instruction Label" (See  
Figure 9.3) on the front panel for the Next Replacement Date of Battery (See the  
technical data prepared by the battery manufacturer).  
10-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
10.3 How to Charge the External Battery  
NOTICE  
The battery is charged while the power is "ON," regardless of the operation state  
of the STP pump.  
The STP control unit is equipped with a constant voltage (55 V) circuit and a constant  
current (0.3 A) circuit for charging the battery.  
After a power failure, charge the battery for approx. 12 times the BATTERY BACKUP time.  
For example, if the BATTERY BACKUP time is 6 minutes, it takes approx. 75 minutes to  
charge the battery.  
If a power failure occurs 2 times within 90 minutes, charge the battery for more than 90  
minutes before starting the STP pump.  
10.4 How to Dispose of the External Battery  
Dispose of or collect worn-out batteries in accordance with the guidelines given by each  
national and/or local government. If you procure the batteries, dispose of them according to  
the guidelines given by the battery manufacturer.  
CAUTION  
Pay attention to short-circuiting or leakage of the batteries when disposing of  
them.  
To prevent the batteries from short-circuiting, remove the lead wires and  
connectors from the batteries.  
Cut the lead wires one by one so as not to cause a short-circuiting.  
Coat the conductive part of the batteries with tape.  
NOTICE  
It is recommended to dispose of worn-out batteries after discharging them  
thoroughly.  
Batteries manufactured by Japan Storage Battery Co., Ltd. will be collected by  
Japan Storage Battery Co., Ltd. For detailed information, contact Japan  
Storage Battery Co., Ltd. or Seiko Seiki.  
10-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
11 Operation Principle of the STP Pump  
The STP-H600/H1000 is a series of magnetically-levitated turbomolecular pumps, featuring  
the following:  
ꢀꢁ Oil free  
ꢀꢁ Low vibration  
ꢀꢁ High reliability  
The STP pump is configured so that rotor blade (1) and stator blade (2) are aligned  
alternately in the axial direction. Gas molecules are pumped from the inlet port to the  
outlet port by the high speed rotation of the rotor.  
By uniting the rotor blades and the sleeve lower, the STP pump has a large capacity  
allowing it to be used in a wide variety of applications. Rotor blade (1) is supported by the  
magnetic bearing without mechanical contact. Therefore, the STP pump requires no  
lubrication oil unlike conventional turbomolecular pumps using ball bearings.  
The magnetic bearing consists of 5 pairs of active magnetic bearings. The rotor is  
supported in the radial direction by 4 pairs of radial direction active magnetic bearings that  
consist of radial sensor (3) and radial electromagnet (4). A pair of axial direction active  
magnetic bearings consists of axial sensor (5) and axial electromagnet (6) to support the  
rotor in the axial direction. Because the rotor is supported without mechanical contact, it  
can rotate at low vibration.  
There is less heat generated from magnetic bearings because there is no friction. Therefore  
the STP pump requires no cooling. However, the STP pump requires water cooling during  
baking and gas suction. Taking into consideration a breakage of magnetic bearings, touch  
down bearings (7) and (8) coated with solid lubrication have been installed. They do not  
contact with the rotor during the rated operation.  
A radial direction sensor, a rotation sensor and a temperature sensor always monitor the  
magnetic bearing as well as the rotor. If an abnormality/error occurs, the rotor will stop.  
11-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
( ) Rotor blade  
1
( ) Stator blade  
2
( ) Touch down bearing  
7
( ) Radial electromagnet  
4
( ) Radial sensor  
3
( ) Touch down bearing  
8
( ) Radial sensor  
3
( ) Radial electromagnet  
4
( ) Axial electromagnet  
6
( ) Axial electromagnet  
6
( ) Axial sensor  
5
Figure 11.1 Cross Sectional View of the STP Pump  
11-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
12 Maintenance and Inspection  
CAUTION  
Stop the STP pump, turn OFF the primary power (switch the breaker "OFF")  
and disconnect the internal battery connector before performing maintenance of  
the STP control unit (battery replacement/fuse replacement).  
Failure to do so may result in an accident such as electric shock.  
Stop the STP pump, and turn OFF the primary power (Switch the breaker  
"OFF") before performing inspections of the STP pump (for deposit).  
Failure to do so may result in the inadvertent rotation of the STP pump, which  
may result in an accident.  
To inspect the air cooling fan and reset the "BATTERY NG" lamp, always be  
sure to stop the STP pump, turn OFF the primary power (Switch the breaker  
"OFF"), open the front panel, then turn ON the primary power. Failure to do so  
may result in electric shock or product damage due to short-circuiting of the  
internal circuit.  
Before performing maintenance or inspections, power OFF the baking heater  
and the air cooling unit.  
DO NOT touch any of the blocks, parts, or connectors inside the STP control  
unit (excluding the battery connector). These parts are excluded from  
maintenance or inspection.  
12.1 Replacing the Internal Battery  
The STP pump is equipped with rechargeable batteries for backup during a power failure  
inside the STP control unit. In order to ensure that the battery functions normally during a  
power failure, replace the batteries before the end of their life (approx. one year).  
See Section 9.5, "How to Replace the Internal Battery."  
The "Battery Instruction Label" is affixed to the front panel. After replacing the battery,  
always record the next replacement date.  
12.2 Inspecting the Air Cooling Fan  
First stop the STP pump. Next turn OFF the primary power (switch the breaker "OFF"),  
open the front panel, and turn ON the primary power (Switch the breaker "ON"). Confirm  
that the air cooling fan (See Figure 4.3) inside the STP control unit is operating. Then, turn  
OFF the primary power (Switch the breaker "OFF") and close the front panel.  
If the air cooling fan is not operating, contact Seiko Instruments to replace it.  
NOTICE  
It is recommended to inspect the air cooling fan when replacing the battery.  
12-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
12.3 Replacing the Fuses  
CAUTION  
Remove the probable causes of the blown fuse (Ex. a problem with the remote  
input signal, etc.) before replacing the fuse.  
Always use specified and rated fuses.  
NOTICE  
If fuses are blown frequently even after they are replaced, contact Seiko  
Instruments.  
The following are protective fuses used in the STP control unit (See Section 4.1.3, "Inside of  
the Control Unit").  
F3, F4: 250 V, 5 A  
For control power protection.  
F5, F6, F7: 250 V, 10 A (arc-extinguishing fuses)  
For I/O TB2 remote input/output terminal protection.  
For power input (19) and (20) and power output (1) to (6) protection.  
F8: 250 V, 10 A (arc-extinguishing fuse)  
For external battery protection.  
F9: 250 V, 10 A (arc-extinguishing fuse)  
For internal battery protection.  
F11, F12: 250 V, 0.1 A  
For START/STOP TB6 remote input signal protection.  
NOTICE  
Two of each three different kinds of fuses are attached for replacement (See  
Table 16.3).  
12-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
12.4 Inspecting for Deposit  
When inspecting for deposit, exhaust gases inside the STP pump thoroughly.  
Residual gases may cause an accident when removing the STP pump.  
Confirm the characteristics of gas to be used, referring to the Material Safety  
Data Sheet (MSDS) you obtain from the gas supplier.  
Wear personal protective equipment if necessary.  
Leaving the STP pump without removing the deposit may cause the STP  
pump to be corroded beyond repair.  
Deposit may have accumulated inside the STP pump depending upon the type of the vacuum  
equipment installed in the STP pump. An increase in the amount of deposit may lead to a  
malfunction of the STP pump.  
Therefore, perform regular maintenance (once every 3 to 6 months as a target).  
Inspect the inside of the outlet port flange for adhesion of deposit while watching the outlet  
port flange. If deposit has accumulated to some extent, an overhaul (cleaning) is required.  
Contact Seiko Instruments.  
Note that the costs of troubleshooting problems resulting from deposit will be at your own  
charge even during the warranty period.  
If a problem resulting from deposit occurs inside the STP pump, hermetically seal inlet port,  
outlet port, purge port, and leak port flanges, and immediately return them to Seiko  
Instruments.  
For the transport method, see Section 12.6, "Transport for Repair or Overhaul."  
Contact with atmospheric air may cause a reaction of the deposit. DO NOT  
open the STP pump to atmospheric air as much as possible.  
Seiko Instruments supplies you the Temperature Management System (TMS)  
unit to keep the temperature high inside the STP pump and prevent the  
accumulation of deposit.  
For details, contact Seiko Instruments.  
12-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
12.5 Overhaul  
The recommended maintenance intervals for different process applications are tabulated  
below:  
Process  
Metal Etch  
Period  
1 year  
Remarks  
Ensure that TMS (Temperature Management  
System) is fitted and operational to prevent  
accumulation of by-product deposition in the  
pump.  
Other Etch  
Processes  
2 years  
It is recommended to change the pump rotor after 5 years due to accumulated wear of  
the protective plating material  
Other semiconductor  
process  
2 years  
Processes resulting in accumulation of deposits  
in the pump will require more frequent service.  
Clean applications  
(Only vacuum pumping)  
Other use  
5 years  
(2 years)  
Dependent on application, contact Seiko  
Instruments or our local Service Center.  
The touch down bearing inside the STP pump will be worn out after a number of  
full speed touch downs.  
The costs of replacing parts that need to be replaced because of deterioration or abrasion will  
be at your own charge.  
When overhaul of the STP pump or the STP control unit is needed, contact Seiko  
Instruments or our local Service Center.  
*1  
Etchingincludes semiconductor etching and LCD etching.  
12-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
STP-H600/H1000 Series Instruction Manual  
12.6 Transporting for Repair or Overhaul  
When removing the STP pump from the vacuum equipment, exhaust gases  
inside the STP pump thoroughly.  
Residual gases may result in an accident when removing the STP pump.  
Confirm the characteristics of gas to be used, referring to the Material Safety  
Data Sheet (MSDS) you obtain from the gas supplier.  
When returning the STP pump which has used corrosive, reactive or  
inflammable gases to Seiko Instruments, specify the type of gas and handling  
precautions on the surface of the package.  
Confirm the characteristics of gas to be used, referring to the Material Safety  
Data Sheet (MSDS) you obtain from the gas supplier.  
When returning the STP pump to Seiko Instruments, be sure to pack it well to  
prevent external damage.  
If any of the specifications has not been satisfied, Seiko Instruments will not be  
responsible for any damage during transport.  
i) Always contact Seiko Instruments before returning the STP pump for repairs, overhaul,  
etc. To avoid any accident by gases or corrosion inside the STP pump, particularly when  
corrosive, reactive or flammable gases have been used, ii) vacuum and hermetically seal the  
STP pump before transport. iii) Specify the type of gases used and handling precautions on  
the package and return the STP pump to Seiko Instruments.  
When returning the STP pump to Seiko Instruments, be sure to pack it well to prevent  
external damage. Use similar or superior packaging material as originally supplied by  
Seiko Instruments.  
The costs of cleaning and overhaul of the STP pump will be at your own charge.  
When returning the STP pump to Seiko Instruments, fill in the necessary items  
in the "STP PUMP PROBLEM CHECK SHEET" and fax it to Seiko Instruments.  
12-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
13 Storage  
13.1 The STP Pump  
When the STP pump is left unused over the specified period (more than a few months),  
follow the precautions below:  
1) Close the inlet port of the STP pump and vacuum it using an auxiliary pump.  
2) Purge dry N2 gas or dry air from the purge port.  
3) Close the outlet port and purge port.  
4) If the STP pump is water cooled, introduce compressed air from one side of the cooling  
water port so that no water remains in the STP pump.  
5) Store the STP pump in a vertical position.  
6) DO NOT store the STP pump in the following places:  
Sꢀ Place of high humidity.  
Sꢀ (If it must be stored in a place of high humidity, insulate it from the outside  
and use a dehumidifying agent.)  
Sꢀ Place of high temperatures (more than + 55 LC) and low temperatures (less  
than - 25 LC).  
Sꢀ Place where there are corrosive gases.  
Sꢀ Place subjected to dripping water.  
Sꢀ Place with a lot of dust.  
Sꢀ Place with insufficient ventilation.  
Sꢀ Place subjected to strong magnetic or electric fields.  
Sꢀ Place subjected to radiation.  
13-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
13.2 The STP Control Unit  
If the STP control unit (including cables) is not used for a long period (more than a few  
months), follow the precautions below:  
1) Remove the internal battery from the STP control unit (see Section 9, "Internal  
Battery").  
NOTICE  
Always remove the battery when the STP control unit is not used for a long  
period.  
2) Store the STP control unit in a horizontal position.  
3) DO NOT store the STP control unit in the following places:  
Sꢀ Place of high humidity.  
(If it must be stored in a place of high humidity, insulate it from the outside  
and use a dehumidifying agent.)  
Sꢀ Place of high temperature (more than + 55 LC) and low temperature (less than -  
25 LC) (excluding the battery)  
NOTICE  
Storage temperatures of the battery range from - 20 to + 40 LC.  
4) Place subjected to direct sunlight.  
5) Place where there are corrosive gases.  
6) Place subjected to dripping water.  
7) Place with a lot of dust.  
8) Place of insufficient ventilation.  
9) Place subjected to strong magnetic or electric field.  
10) Place subjected to radiation.  
11) Place subjected to mist.  
12) Place with electric noise.  
13) Place with vibration.  
13.3 Restarting Precautions  
1) Stored for more than four months:  
Charge internal batteries for 90 minutes or more under the "POWER ON" state before  
starting to operate the STP pump.  
2) Stored for more than one year or when the date of replacement has passed:  
Replace with new batteries.  
13-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
14 Disposal  
14.1 The STP Pump  
Dispose of the STP pump as industrial waste according to the guidelines given by each  
national and/or local government.  
WARNING  
When disposing of the STP pump, exhaust gases inside the STP pump  
thoroughly.  
Residual gases may result in an accident when disposing of the STP pump.  
If the STP pump has been used with reactive or corrosive gasses, always clean  
thoroughly before disposing of it to avoid any injury.  
Confirm the characteristics of gas to be used, referring to the Material Safety  
Data Sheet (MSDS) you obtain from the gas supplier.  
CAUTION  
Seiko Seiki is not responsible for problems during or after disposal.  
14.2 The STP Control Unit  
Dispose of the STP control unit as industrial waste according to the guidelines given by each  
national and/or local government.  
WARNING  
Remove the battery from the STP control unit.  
Failure to do so may result in an accident or fire.  
Always remove the battery from the STP control unit and strictly follow the  
guidelines for battery disposal in accordance with Section 14.3, "How to Dispose  
of The Battery."  
CAUTION  
Seiko Seiki is not responsible for problems during or after disposal.  
14-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
14.3 The Battery  
Dispose of or collect worn-out batteries in accordance with the guidelines given by each  
national and/or local government. If you procure batteries, dispose of them in accordance  
with the guidelines given by the battery manufacturer.  
CAUTION  
Pay attention to short-circuiting or leakage of the battery at their disposal.  
To prevent the battery from short-circuiting, remove lead wires and connectors  
from the battery.  
Cut lead wires one by one so as not to cause short-circuiting.  
Coat the conductive part of the battery with tape or a similar material.  
NOTICE  
It is recommended to dispose of batteries after discharging them thoroughly.  
Batteries manufactured by Japan Storage Battery Co., Ltd. will be collected by  
Japan Storage Battery Co., Ltd. For detailed information, contact Japan  
Battery Co., Ltd. or Seiko Seiki.  
14-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
15 Troubleshooting  
If an abnormality/error is found when using the STP pump, check it and take measures in  
accordance with the following procedures.  
If you cannot trace it or if the STP pump does not function normally after troubleshooting,  
fill in the necessary information in the "STP PUMP PROBLEM CHECK SHEET" and fax  
it to Seiko Seiki.  
CAUTION  
After making sure that the STP pump stops thoroughly, and always turn OFF  
the primary power (Switch the breaker "OFF"), check probable causes and  
remove them.  
15.1 Troubleshooting Immediately After An Abnormality/Error Occurs  
1) In case the “FAILURE” lamp lights due to an overload of the inverter (the STP  
pump continues rotating by inertia: FREE RUN state):  
<MANUAL OPERATION>  
i. Press the MOTOR STOP” switch. The "FAILURE" lamp goes out and the FREE  
RUN state is reset.  
ii. Press the MOTOR START” switch and press the "MOTOR STOP" switch again.  
The brake functions and the STP pump decelerates.  
iii. After the STP pump stops thoroughly (the "BRAKE" lamp goes out: the needle of the  
tachometer is in the red), turn OFF the "POWER ON/OFF" switch, switch the  
breaker "OFF," and remove probable causes of the abnormality/error.  
iv. Restart the STP pump and check if it operates correctly.  
< REMOTE OPERATION>  
i. Slide the "MANUAL/REMOTE" changeover switch on the rear panel to "MANUAL."  
ii. Press the "MOTOR STOP" switch. The "FAILURE" lamp goes out and the FREE  
RUN state is reset.  
iii. Press the "MOTOR START" switch and press the "MOTOR STOP" switch again.  
The brake functions and the STP pump decelerates.  
iv. After the STP pump stops thoroughly (the "BRAKE" lamp goes out: the needle of the  
tachometer is in the red), switch the breaker "OFF," and remove probable causes of  
the abnormality/error.  
v. Restart the STP pump and check if it operates correctly.  
15-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
2) In case the FREE RUN state is not reset after troubleshooting:  
i. Check the tachometer on the front panel.  
ii. If the needle is in the red, it means the STP pump has stopped.  
Power OFF and switch the breaker "OFF."  
iii. If the needle is in the black, the STP pump is rotating.  
Wait for it to enter the red. Then, power OFF and switch the breaker "OFF. "  
iv. Restart the STP pump and check if operates correctly.  
CAUTION  
DO NOT power OFF and switch the breaker OFF until the needle is in the red.  
NOTICE  
If the FREE RUN state cannot be reset, it takes approx. 8 to 10 hours to stop  
the STP pump (the needle is in the red) because it rotates by inertia.  
To stop the STP pump quickly, close the vacuum valve at the outlet port  
flange and introduce gas from the purge port into the STP pump.  
3) Other Cases:  
After the STP pump stops thoroughly, power OFF, switch the breaker "OFF" and remove  
probable causes of the abnormality/error. Then, restart the STP pump and check if it  
operates correctly.  
15.2 Abnormalities When Powering ON  
(When any "Abnormality/Error Warning" lamp lights, see Section 15.5 "When Any  
Abnormality/Error Warning Lamp Lights.")  
Table 15.1 Troubleshooting After Powering ON  
No.  
Symptom  
Probable Cause  
Countermeasures  
Connect the power  
Referred Section  
4.4.2  
The "POWER ON/OFF" (1) Incorrect  
switch built-in lamp  
(green LED) does not  
light, when powering  
ON.  
connection of cable correctly.  
the power  
cable.  
(2) Blown fuses Replace blown fuses 4.1.3  
F3, F4.  
(250 V, 5 A).  
12.3  
(3) Abnormal  
Set the voltage to the 4.4.2  
power voltage  
(4) Power  
failure.  
16.2  
6.1.1  
rated voltage M10%.  
Check whether or  
not a power failure  
has occurred.  
15-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
15.3 Abnormalities When Performing The STP Pump Start Operation  
(When any "Abnormality/Error Warning" lamp lights, see Section 15.5 "When Any  
Abnormality/Error Warning Lamp Lights.")  
Table 15.2 Troubleshooting When Performing The STP Pump Start Operation  
No.  
Symptom  
The "ACCELERATION" (1) TB3  
lamp does not light.  
Probable Cause  
Countermeasures  
When this function 8.2  
INHIBIT (5) - is not used, short the  
Referred Section  
(7) for  
circuit between  
Rotation  
INHIBIT  
terminals (5) - (7).  
If these terminals  
Signal Input are opened, the STP  
are opened. pump will not rotate  
after performing  
start operation.  
(2) The  
Slide the "MANUAL/ 4.1.2  
"MANUAL/R REMOTE"  
5.5  
5.6  
EMOTE"  
changeover switch  
on the rear panel of  
changeover  
switch is not the STP control unit  
in the correct to the correct  
position.  
position.  
(3) Blown fuses Replace the blown  
4.1.3  
F11, 12 (only fuses (250 V, 0.1 A). 8.4  
REMOTE  
12.3  
operation).  
15.4 Abnormalities While the STP Pump is Rotating  
(When any "Abnormality/Error Warning" lamp lights, see Section 15.5 "When Any  
Abnormality/Error Warning Lamp Lights.")  
Table 15.3 Troubleshooting While the STP Pump Is Rotating  
Referred  
Section  
3.3.4  
No.  
Symptom  
Probable Causes  
Countermeasures  
1
The "NORMAL  
OPERATION  
"lamp does not  
light approx. 15  
minutes after  
performing start  
operation.  
(1) Insufficient pressure Use the maximum  
at the inlet port.  
working pressure or less. 16.1  
(2) Insufficient pressure Use the allowable backing 3.3.4  
at the outlet port.  
(3) Leakage of the piping Check piping.  
system.  
(4) Failure in the  
auxiliary pump.  
pressure or less.  
16.1  
3.3.4  
Check the auxiliary pump 3.3  
for its capacity and  
START state.  
16.1  
(use a pump with the  
recommended capacity or  
more).  
(5) Incorrect connection of Connect the motor  
the motor connection connection cable correctly.  
4.4.2  
cable.  
If the cable connector is  
not locked, lock it  
securely.  
15-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
Referred  
Section  
7.3  
No.  
2
Symptom  
Probable Causes  
Countermeasures  
The STP pump  
stops.  
(1) Baking temperature  
overheat.  
Set the temperature to  
120 or lower.  
(2) Insufficient cooling.  
Always cool during baking 7.2  
or gas suction.  
(3) Power failure.  
(4) Over vibration  
(5) (vibration 100 m p-p  
or more).  
Check if power has failed. 6.1.1  
After the STP pump stops 6.1.3  
thoroughly, power OFF  
and remove external  
vibration so as not to  
transmit it to the STP  
pump.  
(6) Overheating inside the Remove heat source or  
STP control unit.  
1) Ambient  
shield for air route.  
1) Keep ambient  
temperature 40 LC or  
lower.  
4.2.1  
temperature  
2) Insufficient  
4.2.2  
12.2  
2) Remove obstacles at  
the ventilation port.  
If the air cooling fan  
is broken, contact  
Seiko Seiki.  
cooling of the STP  
control unit.  
(7) Improper pressure of Use the maximum  
the inlet port.  
(8) Improper pressure of Use the allowable backing 3.3.4  
3.3.4  
working pressure or less. 16.1  
the outlet port.  
pressure or less.  
16.1  
3
Rotor falls during (1) Failure of the STP  
operation. connection cable.  
Inspection is needed.  
Contact Seiko Seiki.  
4.4.2  
(2) Failure of the battery. Inspection is needed.  
Contact Seiko Seiki.  
9
10  
4
Ultimate pressure (1) Failure of the auxiliary Check the auxiliary pump 3.3  
is insufficient.  
pump.  
for its capacity and  
START state.  
16.1  
(use a pump with the  
recommended capacity or  
more).  
(2) Leakage of the piping Check the piping system. 3.3.4  
system.  
(3) Residual molecule.  
If the main composition is 7.3  
H2 or H2O, perform  
baking; if it is other than  
H2 or H2O, clean the  
inside of the vacuum  
equipment  
(If gases remain inside  
the STP pump, contact  
Seiko Seiki when it is  
needed to be cleaned).  
12.4  
15-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
15.5 When Any Abnormality/Error Warning Lamp Lights  
When any Abnormality/Error Warning lamp lights, check which lamp is being lit and see  
Table 15.5 for which procedure No. to follow.  
Then, follow the steps of that procedure No. (shown in Tables 15.6 to 15.8) to remove the  
problem (for the names and positions of the "Abnormality/Error Warning" lamps, see  
Figures 4.1 and 4.3).  
Table 15.4 Cross Reference of Items of Abnormality Warning Lamps  
OVER  
TEMPERATUR  
E
BATTERY NG  
(front panel  
rear side)  
BATTERY  
OPERATION  
EMERGENCY  
OPERATION  
PROCEDURE  
No.  
FAILURE  
No. 1  
No. 2  
No. 3  
No. 4  
No. 5  
No. 6  
(When the  
power is turned  
on again, this  
lamp goes out.)  
:  
The lamp is being lit.  
: The lamp may light.  
—:  
The lamp is not being lit.  
CAUTION  
When checking the " BATTERY NG" lamp, turn OFF the primary power  
(Switch the breaker "OFF"), then open the front panel. After opening the  
front panel, turn ON the primary power (Switch the breaker "ON").  
Failure to do so may result in electric shock or product damage due to short-  
circuiting of the internal circuit.  
NOTICE  
The "BATTERY NG" lamp remains lit until it is reset even after the primary  
power is turned OFF (the breaker is switched "OFF").  
15-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
Table 15.5 Troubleshooting When Any of Abnormality/Error Warning Lamp Lights  
PROCEDU  
RE No.  
No.1  
Referred  
Section  
4.4.2  
Lamp Being Lit  
Probable Causes  
Countermeasures  
OVER  
(1) Incorrect connection of  
the STP connection  
cable.  
Connect cables correctly.  
If the cable connector is not  
locked, lock it securely.  
TEMPERATURE”  
FAILURE”  
EMERGENCY  
OPERATION”  
OVER  
No.2  
(1) Overheat inside the STP  
pump:  
TEMPERATURE”  
FAILURE”  
1) Overheat during  
baking.  
1) Set the temperature of the  
baking heater to 120 LC or  
lower.  
7.3  
2) Insufficient cooling.  
2) At the time of baking or gas 7.2  
suction, always cool the STP  
pump.  
3) Leakage of the  
vacuum piping  
system.  
3) Check the piping system for 3.3.4  
leakage.  
4) Repetitious start/stop 4) Repetitious start/stop  
5.2  
operations.  
operations may cause the  
STP pump to overheat.  
Check a power failure.  
BATTERY  
No.3  
No.4  
(1) Power failure.  
6.1.1  
OPERATION”  
EMERGENCY  
OPERATION”  
FAILURE”  
(1) Incorrect connection of  
the emergency vent.  
valve cable.  
(2) When the rated operation  
does not start approx. 15  
minutes after start  
operation is executed:  
When the  
Connect the cable correctly. If 4.4.2  
the cable connector is not  
locked, lock it securely.  
6.1.4  
ACCELERATION state  
continues for approx. 15  
minutes:  
1) Insufficient pressure at 1) Use the maximum working 3.3.4  
the inlet port.  
pressure or less.  
16.1  
3.3.4  
16.1  
2) Insufficient pressure at 2) Use the allowable backing  
the outlet port.  
3) Leakage of the piping  
system.  
pressure or less.  
3) Check the piping system for 3.3.4  
leakage.  
4) Failure of the auxiliary 4) Check the auxiliary pump  
3.3  
pump.  
for its capacity and START 16.1  
state.  
(use a pump with the  
recommended capacity or  
more).  
5) Incorrect connection of  
the motor connection  
cable.  
5) Connect the cable correctly. 4.4.2  
If the cable is not locked,  
lock it securely.  
15-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
PROCEDU  
RE No.  
No.4  
Referred  
Section  
Lamp Being Lit  
Probable Causes  
Countermeasures  
FAILURE”  
(3) Overheating inside the  
STP control unit:  
1) Ambient temperature. 1) Set the ambient  
temperature to 40 LC or  
4.2.1  
lower.  
2) Insufficient cooling of 2) Remove obstacle from the  
4.2.2  
12.2  
8.2  
the STP control unit.  
ventilation port.  
If the air cooling fan is  
broken, contact Seiko Seiki.  
Remove abnormalities from  
peripheral equipment.  
(4) Peripheral equipment  
abnormality (when  
peripheral equipment is  
connected between TB3  
R.P. THER (1)–(3)).  
(5) TB3 R. P. THER (1)–(3) Short the circuit between  
8.2  
are opened.  
terminals (1)–(3).  
(when peripheral  
equipment is not  
connected between TB3  
R. P. THER (1)–(3).  
(6) Insufficient charging or Check that the battery is fully  
9
deterioration of the  
battery.  
charged. If the battery is not  
charged, charge it. Check the  
battery for deterioration and  
the next replacement date.  
If the battery deteriorates or  
its replacement date has  
passed, replace it.  
10  
(7) Incorrect connection of  
the battery.  
Check that external or internal 9  
batteries are correctly  
connected.  
10  
(8) Blown fuses F8, F9 (for  
battery protection).  
Replace the blown fuses  
(250 V, 10 A).  
12.3  
EMERGENCY  
OPERATION”  
No.5  
(1) Different model name or Check that the model name,  
Safety  
No. on the name plate.  
serial number and cable length Precautions  
of the STP pump, the STP  
control unit and the STP  
connection cable are the same.  
After the STP pump stops  
thoroughly, power OFF and  
remove external vibration so  
as not to transmit it to the STP  
pump.  
4.4.2  
(2) Excessive external  
vibration (vibration of  
100 m p-p or more).  
6.1.3  
15-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
PROCEDU  
RE No.  
No.6  
Referred  
Section  
Lamp Being Lit  
Probable Causes  
Countermeasures  
FAILURE”  
(1) Insufficient charge or  
deterioration of the  
battery.  
1) Check whether or not the  
9
BATTERY NG”  
(inside the front  
panel)  
( “EMERGENCY  
OPERATION”  
lamp may light)  
battery is fully charged. If 10  
the battery is not  
sufficiently charged, charge  
it.  
2) Check the deterioration and  
the replacement date of the  
battery. If the battery has  
deteriorated or the  
replacement date has  
passed, replace the battery  
with a new one.  
3) After taking steps 1) and 2)  
above, reset BATTERY NG.  
CAUTION  
When checking the " BATTERY NG" lamp, turn OFF the primary power  
(Switch the breaker "OFF"), then open the front panel. After opening the  
front panel, turn ON the primary power (Switch the breaker "ON"). Failure to  
do so may result in electric shock of product damage due to short-circuiting  
of the internal circuit.  
NOTICE  
The "BATTERY NG" lamp remains lit until it is reset even after the primary  
power is turned OFF (the breaker is switched "OFF").  
15-8  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
16 Specifications and Accessories  
16.1 Specifications for the STP Pump  
Table 16.1 Specifications for the STP Pump  
STP-H600  
Item  
STP-H1000  
Inlet port flange  
ICF203/VG150/ISO160 ICF253/VG200/ISO200  
KF40  
Flange size*1  
Outlet port flange  
N2  
He  
N2  
N2  
He  
N2  
L/s  
L/s  
L/s  
600  
550  
500  
1000  
900  
800  
Pumping speed  
>108  
104  
Compression ratio  
Ultimate pressure  
103  
Pa (Torr)  
10-7(10-9) order (after baking)  
67 (0.5)  
Maximum working When cooled water:  
pressure  
When cooled air:  
When cooled natural  
air:  
When TMS unit is  
used:  
5.3 (0.04)  
Pa (Torr)  
0.9 (0.007)  
13(0.1)  
Allowable backing When cooled water:  
pressure  
When TMS unit is  
Pa (Torr) used:  
400 (3)  
267 (2)  
When cooled air:  
When cooled natural  
air:  
Rated speed  
rpm  
min  
min  
μm O-P  
dB  
35,000  
Starting time  
Stopping time  
Vibration  
6
6
0.01 (at 35,000 rpm)  
Noise  
50 (at 35,000 rpm)  
Baking temperature  
Lubricating oil  
Installation position  
Cooling method*2  
Recommended auxiliary pump L/min  
Mass*3  
LC  
120  
Not necessary  
Free  
Water cooled, Air cooled, Natural air cooled  
240  
kg  
LC  
LC  
31  
32  
Ambient temperature range  
Storage temperature range  
0 to 40  
- 25 to + 55  
Values shown in the above table are typical. They are not guaranteed.  
*1 : ICF (JVIS) VG, KF (JIS)  
*2 : Varies depending upon the operation pressure range.  
*3 : Mass is a value of state that the only standard accessory was installed. (except the optional accessory)  
16-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
STP-H600/H1000 Series Instruction Manual  
16.2 16.2 Specifications for the STP Control Unit  
Table 16.2 Specifications for the STP Control Unit  
Item  
Specifications  
200 to 240±10%  
2.6  
Input voltage  
Input power  
V AC  
kVA  
kVA  
Hz  
At start:  
Under rated operation:  
1.0 to 2.6  
50/60±2  
Input frequency  
Input phase  
Rated current of main breaker  
Motor drive system  
Single phase  
20  
3-phase transistor-inverter  
A
Output voltage under rated  
operation(3-phase)  
Output frequency under rated  
operation  
V AC  
Hz  
52  
584  
Ambient temperature range  
kg  
0 to 40  
Storage temperature range  
- 25 to + 55 (excluding a battery)  
Battery storage temperature range  
- 20 to + 40  
40  
Mass  
Battery used  
12 V 2Ah x 4 (J apan Storage Battery Co.,  
Ltd./Small-Sized Sealed Lead-Acid  
Battery/PORTALAC series PE12V2WS)  
[Front Panel]  
Operation switches  
Sꢀ POWER ON/OFF  
(Illuminated alternate push button  
switch, black/green)  
Sꢀ MOTOR START  
(Momentary push button switch, black)  
Sꢀ MOTOR STOP  
(Momentary push button switch, red)  
Sꢀ HEATING ON/OFF  
(Illuminated alternate push button  
switch, black/green)  
[Rear Panel]  
Sꢀ MANUAL/REMOTE Changeover  
(Toggle switch)  
Panel indication lamps  
Sꢀ ACCELERATION  
Sꢀ NORMAL OPERATION (Green LED)  
Sꢀ BRAKE (Yellow LED)  
Sꢀ EMERGENCY OPERATION (Red LED)  
(Green LED)  
Sꢀ FAILURE  
(Red LED)  
(Red LED)  
(Red LED)  
Sꢀ OVER TEMPERATURE  
Sꢀ BATTERY OPERATION  
Sꢀ [Switch built-in Lamps]  
Sꢀ POWER ON (POWER ON/OFF, green)  
Sꢀ HEATING ON (HEATING ON/OFF,  
green)  
*1 : Mass is a value of state that the only standard accessory was installed. (except the optional accessory)  
16-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
STP-H600/H1000 Series Instruction Manual  
Item  
Input/output terminal  
Specifications  
Sꢀ AC POWER TB1 (3 pins)  
Sꢀ P.CONNECTOR CON5A (45 pins)  
Sꢀ INVERTER OUTPUT CON2A (5 pins)  
Sꢀ LEAK VALVE CON8A (8 pins)  
Sꢀ I/O TB2 (20 pins)  
Sꢀ DC I/O TB3 (10 pins)  
Sꢀ START STOP TB5 (3 pins)  
Sꢀ START/STOP TB6 (2 pins)  
Sꢀ EXT. BATTERY CON9 (2 pins)  
Sꢀ Primary power cable  
Input/output cables  
Safety functions  
Sꢀ STP connection cable  
Sꢀ Motor connection cable  
Sꢀ Emergency vent. valve cable  
Sꢀ Ground cable  
Sꢀ Magnetic bearing abnormality/error  
detection  
(Rotor vibration: 100μm p-p or more)  
Sꢀ STP pump overheat detection  
(Motor: 110 over)  
Sꢀ Inverter overload detection  
(Inverter current: 40 A over,  
instantaneous, A over for 15 minutes)  
Sꢀ Battery abnormality/error detection  
(Battery terminal voltage: 42 V under)  
Sꢀ Power failure detection  
(Detects power failure of over about 2  
seconds)  
Sꢀ STP pump overspeed detection  
(37500 rpm over)  
Sꢀ STP control unit overheat detection  
(90over)  
Sꢀ Brake failure detection  
16-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
16.3 Accessories  
Table 16.3 Accessories  
Item  
Q'ty  
1
Remarks  
STP control unit  
Primary power cable*1  
STP connection cable*1  
Motor connection cable*1  
Ground cable*1  
1
1
1
1
Inlet port cover  
1
Outlet port cover  
1
STP connector cover  
Motor connector cover  
Leg  
1
1
8
4 legs are attached to the STP pump  
Rubber foot for lug  
4
1
1
1
1
2
2
1
Emergency vent. valve  
Emergency vent. valve clamping ring  
Emergency vent. valve O-ring washer  
Blank flange for purge port  
Purge port clamping ring  
Purge port O-ring washer  
T-piping  
With a one-sided connector cable*1  
KF*210  
KF*210  
KF*210  
KF*210  
KF*210  
KF*210  
External battery connection plug  
Spare fuses 250 V, 5 A  
250 V, 10 A  
1
2
2
F3, 4  
F5, 6, 7 ,8 ,9  
(arc-extinguishing fuses)  
250 V, 0.1 A  
Instruction Manual  
2
1
F11, 12  
16.4 Recommended Spare Parts  
Table 16.4 Recommended Spare Parts  
Item  
Model and Manufacturer  
Specifications  
Fuses for F3, 4  
MF60-NR5, 5 A, 250 V  
(TOYO FUSE CO., LTD)  
5 A, 250 V, Normal acting fuse  
6.35 × 30 mm  
Fuses for F5 to 9  
MF60-NR10-S, 10 A, 250 V  
(TOYO FUSE CO., LTD)  
10 A, 250 V, Normal acting fuse  
(arc-extinguishing fuses)  
6.35 × 30 mm  
Fuses for F11, 12 MF60-NR0.1, 0.1 A, 250 V  
(TOYO FUSE CO., LTD)  
0.1A, 250 V, Normal acting fuse  
6.35 × 30 mm  
Internal batteries PE12V2WS  
Nominal voltage: 12 V  
(J apan Storage Batteries Co., Ltd.) Rated capacity (20-hour rate): 2Ah  
CAUTION  
Do not use parts other than those recommended in the above table.  
If you use these parts, contact Seiko Seiki before use.  
*1 :The standard cable length is 5 m.  
*2 :JIS  
16-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
STP-H600/H1000 Series Instruction Manual  
5 .  
1 4 7  
6 0  
5 .  
1 4 1  
3 0  
5  
1
5
3
5 0  
3 5  
1 3  
3 9  
2 6  
8 0  
1
Figure 16.1 External Appearance of the STP Pump (Example: STP-H1000 Series)  
16-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
2 4  
1 8 8 . 6  
5
. 5 3 4  
5 . 4  
1 3 0  
1 9 9  
3 4 . 5  
e ) a c p S e l b a ( C  
5 0 ) ( 1  
5 0 0  
4 7 0  
3
3 0  
5 0  
5 0  
5 0  
2 4 . 5  
. 5 2 4  
6
3 5 4  
Figure 16.2 External Appearance of the STP Control Unit  
16-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
Figure 16.3 Label Affixing Positions for the STP Pump  
16-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
. l  
p a t n e o r n f f o d i e s n i h t e o t l  
a l b s e i h t x i f f A  
.
y n o l l  
o r n i e a d r t i k i e S o k i e S b y  
d e m o f r p e b e l d u o h s n i u n T  
o s n r n e e p e z d i h t o r a u  
n . i o u t a c  
h t i w e l d n H a  
注 意 取 扱  
k g  
g
N O I U T C A  
k g  
て く だ さ い 許 。 可 を 受 け  
の ト レ ー 精 ニ 機 ン  
整 を 行 浮 う 上 場 調 合  
た は グ ま  
イ コ は セ  
物 重 量  
意 ・ 注  
t
d u o r c p y e a H v  
・ 注 意  
C A U T  
N O I  
N e x  
交 換 日 次 回  
e t a d  
n a g h e c e x t  
.
s e i r e t b a t  
い 。  
リ ー  
な い で く だ さ 同 時 を に 使 用 し  
と 外 部 バ ッ テ 部 バ ッ ◇ テ 内 リ ー  
l
n a e r n i t h i t w n o i i n b a t m o c i n  
e i r s e t b a t n r a e l t e x e s u n o t D o  
さ い だ 。  
.
r
C h a n g e  
て く  
年 毎 に 交 換 し ッ テ リ ◇ ー バ は 1  
e y a  
y
e v r e e i r s e t b a t  
・ 注 意  
N O I U T C A  
SAFETY INSTRUCTIONS  
Read instruction manual  
安全上の指示事項  
使用前に取扱説明書を  
お読みください。  
before operation  
.
CAUTION  
Use pump control unit and  
・注意  
◇ポンプとコントロールユニッ  
トおよび接続ケーブルは同一  
№の組合せで使用してくださ  
い。  
◇必ず指定の型式のケーブルを  
使用してください。  
◇最初にアース線を接続してく  
ださい。  
,
connection cable with same  
CAUTION  
Lock connector  
・注意  
◇コネクタは確実に  
ロックしてくださ  
い。  
◇ブレーカONの時  
はコネクタを外さ  
ないでください。  
Lock  
serial number  
.
securely  
.
Use only designated type of  
Do not remove  
connector while  
breaker is ON  
cables  
Connect ground cable first  
.
NO  
.
.
200 240V  
From viewing arrow A  
TURBOMOLECULAR PUMP CONTROL UNIT  
MODEL:  
MANUFACTURING DATE:  
POWER SUPPLY( Hz):  
SERIAL NUMBER:  
POWER CONSUMPTION:  
VOLTS WEIGHT:  
VA  
kg  
STP-  
No.  
/
50 60 1φ  
MANUFACTURED BY: SEIKO SEIKI CO LTD YASHIKI NARASHINO CHIBA JAPAN  
., . ,  
LENGTH  
M
SEIKO SEIKI  
CE  
Figure 16.4 Label Affixing Positions for the STP Control Unit  
16-8  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP-H600/H1000 Series Instruction Manual  
・警告  
WARNING  
Hot surface  
.
Can burn hands  
Do not touch  
高温部  
.
やけどします。  
触らないでく  
ださい。  
.
Baking Heater  
Air Cooling Unit  
AIR COOLING UNIT  
VOLTS AC  
SEIKO SEIKI MADE IN JAPAN  
V
Figure 16.5 Label Affixing Positions for the Special Accessory  
Figure 16.5 Label Affixing Positions for the Special Accessory  
16-9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STP PUMP PROBLEM CHECK SHEET  
Contact your nearest Seiko Seiki’s office shown in the rear cover.  
To : Company Name  
Contact Name  
TEL. Number  
FAX. Number  
Address  
:
:
:
:
:
Date :  
From: Customer Name  
Person in charge  
TEL. Number  
:
:
:
:
:
Date :  
FAX. Number  
Address  
Input Voltage:  
Model Name :  
STP-  
Inlet Flange Type :  
Length of Connection Cable:  
V
m
M/C No. :  
Manufacturing Date :  
Other:  
STATE OF PROBLEM  
Check Items  
Abnormality Warning  
(Check lamps being lit)  
State/Result  
OVER TEMPERATURE  
BATTERY OPERATION  
FAILURE  
(while the vacuum chamber is in operation, etc.)  
OTHER INFORMATION  
EMERGENCY  
OPERATION  
BATTERY NG (inside  
panel)  
Yes / No  
Pump  
Noise  
Yes / No  
Vibration  
Heat  
Yes / No  
Yes / No  
Choke Water  
Deposition  
Yes / No  
Control Unit Next Battery  
Replacement Date  
YY  
/
MM  
Power Failure  
Time  
Min.  
Blown Fuses  
Yes / No  
(Fuse No. F  
)
Yes / No  
Heat  
Other  
Used Gas  
Pump Installation  
Position  
Vertically / Horizontal  
Upside-down  
Other  
(
)
Yes / No  
Leakage  
TMS Unit  
Yes / No  
Seiko Seiki Comment :  
Date  
.
.
Name  
Job No.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
UNIT CONVERSION TABLE  
Length  
m
cm  
mm  
inch  
3
1.00x10  
10.0  
1
1
100  
1
39.4  
0.394  
39.4x10  
1
0.01  
1x10  
-3  
-3  
0.10  
2.54  
-3  
25.4x10  
25.4  
Mass  
g
1
kg  
lb.  
-3  
-5  
1.00x10  
1
2.20x10  
2.20  
1
3
1x10  
454  
0.454  
Pressure  
2
kgf/cm  
Pa  
1
Torr  
7.50x10  
1
-3  
-5  
-3  
1.02x10  
1.36x10  
1
133  
4
9.81 x 10  
736  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Sales Office (Japan)  
Seiko Instruments Inc. Vacuum Innovative Products Sales Department  
1-8 Nakase, Mihama-ku, Chiba-shi, Chiba 261-8507 JAPAN  
Telephone: <Vacuum Innovative Products Sales Department, Direct>  
Domestic  
043-211-1401  
International  
Domestic  
+ 81-43-211-1401  
043-211-8075  
Facsimile:  
International  
+ 81-43-211-8075  
Manufacturer, Sales Office (Japan)  
Seiko Instruments Inc. Vacuum Innovative Products Division  
4-3-1 Yashiki, Narashino-shi, Chiba 275-0004 JAPAN  
Telephone: <Vacuum Innovative Products Service Section, Direct>  
Domestic  
047-475-3123  
International  
+81-47-475-3123  
<Company main>  
Domestic  
047-475-3111  
International  
+81-47-475-3111  
Facsimile:  
Domestic  
047-471-1544  
International  
+81-47-471-1544  
Supplier:  
Seiko Instruments Inc.  
1-8 Nakase, Mihama-ku, Chiba-shi, Chiba 261-8507 JAPAN  
Telephone:  
national 043-211-1401 (VI Sales Sec., Direct)  
intern. + 81-43-211-1401  
Facsimile:  
national 043-211-8075  
intern. + 81-43-211-8075  
Supplier:  
Seiko Instruments USA  
1130 Ringwood Court  
San Jose, CA 95131 U.S.A.  
Telephone:  
national 408-922-5932  
intern. +1-408-922-5932  
national 408-922-1959  
intern. +1-408-922-1959  
Facsimile:  
Supplier:  
Seiko Instruments Belgium S.A  
Av. De Messidor 202 1180 Bruxelles, Belgium  
Telephone:  
national 02-346-6274  
intern. + 32-2-346-6274  
national 02-347-5268  
intern. + 32-2-347-5268  
Facsimile:  
Download from Www.Somanuals.com. All Manuals Search And Download.  

PYLE Audio Stereo System PLCD20CH User Guide
Pyramid Car Audio Car Stereo System CDR 36DX User Guide
Realistic Scanner PRO 39 User Guide
Rival Gas Grill GRF405 User Guide
Ryobi Outdoor Trimmer RPT2543C, RY7011 User Guide
Samsung Stereo System MAX ZL65 User Guide
Samsung Vacuum Cleaner VCC96P0H1G User Guide
Sanyo Heat Pump APR CHRZP900BAB Kit User Guide
Sears Air Compressor 919727321 User Guide
Senco Nail Gun SFN2 User Guide