Philips Microscope Magnifier CM 20 User Manual

OPERATION PROCEDURES FOR PHILIPS/FEI CM-20  
TRANSMISSION ELECTRON MICROSCOPE  
General Comments  
Normally the microscope will be left in the power ON (standby ready to operate)  
condition; i.e., all circuits will be continuously energized and the microscope will be under full  
vacuum (MAIN, VACUUM, HIGH TENSION lamps on the right control panel are all lit if the  
PANEL DIM is pressed). If it is not in this condition do not operate the microscope. Seek help  
from authorized personnel.  
If the microscope is in standby ready condition, proceed with the following abbreviated  
procedures for high voltage application and alignment before beginning any observation. If the  
microscope is not properly aligned, good results will not be obtained.  
I. Start-Up (This procedure is when the microscope starts from cold)  
(1) Run the cooling water (for vacuum and lens systems).  
(2) Turn on the main power line (in mechanical room, behind the microscope room).  
(3) Press the ON power switch on MICROSCOPE. The microscope will go through a series  
of self check by the computer program. It will automatically turn on VACUUM  
SYSTEM, and reach to the high vacuum state as indicated by the UHV and HI VAC  
lights.  
II. Specimen Exchange  
(i) Specimen Unloading:  
(1) Make sure specimen position is at center, i.e., SHIFT (right at 0, left at 10), TILT  
indicator at 0 and 11 (HOLDER/azimuth) and tilt drive is engaged.  
(2) From holding position (B): Pull out the specimen holder gently until it no longer moves  
and then turn it clockwise about 180 degree until it stops (A position, make sure it no  
longer turns, the indicator box face down).  
(3) Gently extract the holder from the airlock.  
(4) Place the specimen holder onto a proper sample holder stand/support for specimen  
loading/unloading.  
(ii) Specimen Loading:  
(1) Set specimen onto suitable specimen holder. USE GLOVES! Caution: No direct  
contact of your hand with a specimen holder and any other part which will be placed  
inside the microscope.  
WAC, 09/27/02  
1
Download from Www.Somanuals.com. All Manuals Search And Download.  
(4) Make sure the camera cover is correctly placed, and then press CAM AIR again to pump  
the camera chamber. You should be able to hear the mechanical pump starting. Pull on  
the camera lid to ensure the vacuum is pulling.  
(5) Press READY, then TEM CAMERA, and CAM INIT.  
(6) Press RESET to reset film number (to 36), and press READY.  
(7) Press MODES and then READY to go to start-up page.  
(8) Develop your films (negatives). Be careful not to contaminate any chemicals onto the  
unexposed films when you changing films in the dark room. Make sure there are 36  
films in the unexposed camera box.  
(9) On the start-up page, if the VACUUM STATUS says READY, push HIGH TENSION  
button and the green LED should light. If the VACUUM STATUS says START-UP (i.e.,  
the microscope column not reach to high vacuum yet), wait for it to say READY and then  
push the HIGH TENSION button.  
(10) Pull the PANEL DIM knob and leave the microscope for the next user.  
(11) Fill out TEM log book.  
(12) Make sure the microscope room and working areas (TEM consol, table/bench, and put  
chair into proper position…etc.) are clean before you leave.  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
ROUTINE ALIGNMENT PROCEDURES FOR PHILIPS CM-20 TEM  
The following protocol is not intended to be specific. It is intended to be used as a mean  
to refresh one’s memory in carry out step by step procedures which are necessary to obtain  
reasonably good results with the transmission electron microscopes. More specific information  
can be obtained from the instruction manual from the Microscopist in Charge or the Center  
Director. Please do not hesitate to ask about anything you do not understand.  
I. Before Starting, Verify:  
(1) Check log book. Enter your name in the log book before any action on the microscope.  
Note cumulative hours on filament and past difficulties. Note: You must have your name  
on the reservation.  
(2) Push PANEL DIM and turn DATA DIM knobs in to turn on display.  
(3) Ascertain vacuum in good condition by press VACUUM on the screen. IGP<25. UHV,  
HIVAC (under vacuum system) are lit.  
(4) Fill cold trap with liquid nitrogen (LN2 cylinder located outside the TEM room).  
(5) All apertures out (levers should be to the right) except condenser aperture.  
(6) Tilt indicator is at 0o (TILT) and 11o (HOLDER) degree, respectively.  
(7) Specimen holder position at 0 (right) and 10 (left) position.  
II. To Apply High Voltage:  
Normally the microscope is in standby condition with HV on. Check high tension LED  
(green light). If it is on, go to step III. If it is not, you need to turn on the voltage.  
(1) Press MODES and TEM to enter the TEM bright field page.  
(2) Press PARAMETERS on the TEM bright field page.  
(3) Select Emission 1 (or the default number) and High Tension 20 kV using soft keys. If the  
microscope has been used, then set at 200 kV (see below),  
(4) Make sure IGP<25 on vacuum page.  
(5) The following procedures (step 5 through 8) are applied after changing a new filament or  
the accelerating voltage has been off for some time (a few days). If the microscope has  
been continuously used, then make sure the setting is at 200 and push HIGH TENSION to  
turn on the accelerating voltage. Go to step III (to apply filament current) after the HT  
has stabilized (i.e., Emission meter is stable).  
4
Download from Www.Somanuals.com. All Manuals Search And Download.  
(6) Set voltage at 20. Push HIGH TENSION button (the green LED should light)  
Slowly increase voltage from 20 kV to 160 kV using soft keys (wait 1 minute between  
each step).  
(7) After voltage reaches 160 kV, press FREE CONTROL soft key. Use focus knob, slowly  
increase voltage from 160kV to 200kV. If it arcs (loud popping sound), use focus step size  
knob to decrease voltage increment. You may also increase voltage to 200 kV by directly  
press 200 if the microscope has been conditioned and continuously used.  
(8) After 200 kV is reached, press FREE CONTROL soft key to un-highlight it.  
(9) Press READY (button under computer display) to go back to TEM bright field page.  
III. To Apply Filament Current:  
(1) Press READY to go back to start-up page (or TEM bright field page if high tension was  
originally off).  
(2) Press MODES on start-up page (or TEM bright field page)  
(3) Press CONFIGURATION on mode selection page.  
(4) Turn FILAMENT knob slowly (with 10-20 seconds pause between each setting/click) until  
the number after ACTUAL is equal to the number after FIL LIMIT. DO NOT CHANGE  
FILAMENT CURRENT AND EMISSION SETTINGS WITHOUT  
AUTHORIZATION.  
(5) Check EMISSION current. It should be somewhere between 5 to 8 µA.  
(6) You should be able to see the intensity on the fluorescent screen. If you are at a  
magnification >10,000, you will need to condense the beam in order to see the intensity.  
(7) Press READY to go back to mode selection page and then press TEM to enter TEM bright  
field page.  
IV. Manual Axial Alignment of the Illumination System  
(A) Centering Condenser Aperture  
(1) Select SPOT SIZE 2, set magnification at 15,000X to 25,000X.  
(2) Move your sample to the center hole (i.e., away from the sample).  
(3) Select condenser aperture by carefully turning largest knob on end of aperture (if needed).  
(4) Center the condenser aperture. Condense the beam using the INTENSITY knob and  
center the beam, as necessary, using the SHIFT knobs. Aperture center can be judged  
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
by first making crossover image with INTENSITY knob, then open up condenser II  
(INTENSITY)/expand the beam and close condenser II on the opposite side of crossover.  
If the beam appears to swing this indicates the aperture is not properly centered. Center  
the aperture by moving the X and Y controls on the aperture housing.  
(5) Repeat the last step until the illuminated area remains centered and becomes  
concentrically smaller or larger on changing the setting of the INTENSITY knob.  
(B) Center Filament Image and Alignment of Condenser System:  
(1) Under-saturate filament by turning FILAMENT knob counter-clockwise slowly, at  
crossover, until filament image is seen (characteristic “donut” shape). You may also turn  
the FILAMENT to approximately 4-5 steps below preset saturation point to observe the  
filament image.  
(2) Gun tilt/shift alignment  
- Press ALGN to activate alignment selection.  
- Press GUN TILT soft key.  
- Condense beam.  
- Slowly turn MULTIFUNCTION X and Y knobs to get maximum illumination, or center  
the filament image.  
- Press GUN SHIFT soft key in ALGN  
- Use MULTIFUNCTION X and Y to center the beam.  
(3) Alignment of condenser system  
- Filament remains under-saturated.  
- ALGN key remain at GUN SHIFT.  
- At SPOT SIZE 2 (large spot size), condense beam, center the beam with  
MULTIFUNCTION X and Y.  
- Go SPOT SIZE 9 (small spot size), condense beam, center the beam with beam SHIFT  
X and Y knobs.  
- Go to SPOT SIZE 2, and repeat the above two steps, until there is no significant  
deviation when change SPOT SIZE between 2 and 9.  
- Press ALGN to deactivate alignment selection.  
6
Download from Www.Somanuals.com. All Manuals Search And Download.  
(4) Condenser lens astigmatism correction  
- Press STIG to activate stigmator control.  
- Press COND soft key so that it is highlighted (the default is OBJ).  
- Condense the beam.  
- Use MULTIFUNCTION knobs to correct the astigmatism. The beam should be round  
when changing the setting of the INTENSITY knob, or the filament image will be  
sharpest if filament is under-saturated.  
- To see the sharp image of filament, three conditions, under-saturated, crossover, and  
free from astigmatism, must be met.  
- Press STIG to de-activate stigmator control.  
(4) Pay attention to filament image while slowly increase filament current to saturate (prefer  
slightly under-saturate) position. Emission current should be 5~8 µA.  
(5) Check aperture center by spreading the beam, make sure that beam spreads without sweep.  
(C) Alignment of Imaging System:  
(i)  
Voltage Center Alignment  
To align the imaging system, the AC component is superimposed on the accelerating voltage  
in order to move the image radially. The center of the radial movement (called “voltage  
center”) is then matched with the center of the fluorescent screen. The main purpose of this  
step is to center the beam tilt so there won’t be a tendency to have a two-dimensional image  
smeared out in one direction. This is very important in doing high-resolution work.  
(1) Z-alignment (eucentric height adjustment)  
- Rough focus the sample at 10,000 to 20,000X.  
- Find an identifiable object and center it on the screen  
- Tilt the sample. This may cause the center of the field of view to move towards the edge  
of the projection screen if the specimen is not at eucentric height position.  
- Set the specimen height at the point where there is no movement of the specimen upon  
tilting by adjusting the z control (screw on the goniometer just below the sample holder).  
- Bring the tilt mechanism to 0o.  
(Hint: if autofocus is way off, z-alignment is bad. Focus the image with the z-alignment  
from autofocus (use screw on the goniometer just below the sample stage))  
(2) Pivot point X and Y  
- Move to center of sample (i.e., any hole or away from the sample).  
- Press PIVOT PT X soft key in ALGN.  
- Condense beam.  
- Use MULTIFUNCTION knobs to bring two circles together.  
- Press PIVOT PT Y soft key in ALGN.  
- Condense beam.  
- Use MULTIFUNCTION knobs to bring two circles together.  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
- Press ALGN to de-activate.  
(3) Voltage centering  
- Place a more or less symmetrical image to the center.  
- Condense the beam at the center of screen, and spread the beam.  
- Press ROT CENTER and VOLT soft keys in ALGN.  
- Make sure it is in the VOLT mode (check bottom right soft key).  
- Determine the voltage center from the movement of the entire image and match the  
center of image movement with that of the fluorescent screen by adjusting  
MULTIFUNCTION knobs.  
- If necessary, increase focus step to increase amplitude to see image movement  
- Press ALGN to stop.  
- Condense the beam at the center and spread the beam to check.  
(4) Repeat steps 4 at higher magnification  
(5) Press ALGN to deactivate alignment selection.  
(ii)  
Current Center Alignment (Option)  
The image on the screen rotates when focus is changed by changing the objective lens exciting  
current. The center, of this rotation is called “Current Center”, and should more or less  
coincide with the screen center. If the current center does not coincide with the screen center,  
it is difficult to bring the image quickly into focus, not to mention the possible adverse effect  
on microscope performance.  
However, there is no need for current centering if voltage center is properly aligned. Perform  
the current center alignment, using CURRENT mode of ROT CENTER soft key in ALGN to  
place the center of image spin in the center of screen with MULTIFUNCYION knobs.  
(iii)  
Intermediate and Project Lens Alignment  
(1) Center and fully spread the beam.  
(2) Depress D switch to obtain a diffraction pattern.  
(3) Press DIF ALIGNM soft key on ALGN (alignment) page.  
(4) Align the diffraction pattern using MULTIFUNCTION knobs at camera length 4.40 m.  
8
Download from Www.Somanuals.com. All Manuals Search And Download.  
(5) Change camera length (by changing MAGNIFICATION knob) to 44.0 mm.  
(6) Realign the diffraction pattern.  
(7) Repeat steps 4 to 6 until the transmitted spot remains at (near) the center while changing  
camera length between 44.0 mm to 3.2 m.  
(8) Deactivate ALGN, and depress D to get out of diffraction mode.  
(iv) Insertion and centering objective aperture  
(1) Move to the edge of the sample.  
(2) Make sure aperture is out.  
(3) Press D to get to diffraction mode.  
(4) Put objective aperture in (lever to the left)  
(5) Select aperture size (the second largest, 40 microns, is usually good)  
(6) Center aperture around transmitted beam using knobs on aperture  
(7) Press D to get out of diffraction mode.  
(D)Astigmatism Correction of Objective Lens (section 2.5.4 in TEM manual):  
Astigmatism of objective lens appears as an image blurring phenomenon which one  
image is accompanied by another different-focus image at right angles to each other. Namely,  
because of asymmetrical focus, an image does not become sharp and appears to drift in one  
direction. Under this condition, exact focusing is impossible. Hence, for recording a quality  
micrograph, it is indispensable to correct astigmatism of the objective lens.  
Three techniques (for correcting objective lens astigmatism) are commonly used depend  
on the magnification and material.  
First Method (for low magnification images, <100kx)  
(1) Find a suitable size of hole in a holey carbon or any sample.  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
(2) Set a magnification at least 3 or 4 times higher than that will used for photographing.  
(3) Obtain the thinnest over-focus (Fresnel) fringe that can be clearly seen in the binocular,  
with beam spread. Caution: make sure you focus the binocular on the viewing screen.  
(4) Adjust the OJB STIG X and Y (MULTIFUNCTION keys) until the fringe follows the  
outline of the hole uniformly and symmetrically.  
(5) Use finest focus step, go through focus series to check astigmatism correction.  
Second Method (for specimens with few small holes or any specimen)  
(1) Using any specimen, adjust (at the best focus) objective stigmators until image is as crisp  
(sharp) as possible.  
(2) Go through focus series to check astigmatism correction.  
Third Method (for ultra-high resolution images, >150kx )  
(1) Find the amorphous area such as amorphous carbon film or any amorphous contaminants  
around the specimen or any amorphous material.  
(2) If there is an appearance of smearing or some thing that looks like periodic structure  
exists in the amorphous film, this is an indication of an astigmated objective lens. Correct  
this by observing the amorphous image as the objective lens focus is changed.  
(3) Correct astigmatism with the OBJ STIG X and Y (MULTIFUNCTION) controls. There  
should be random grains in the structure/image of the carbon or amorphous film.  
(4) Go through finest focus series to check astigmatism correction. It should have no more  
“directional” on the image when changing focus.  
Note: Be very careful about correcting the objective astigmatism. If it looks pretty good, you  
may want to leave it rather than risk making it worse if you were not sure to make a better  
correction. The objective astigmatism is most critical at high magnification imaging.  
Final Comments  
You are now ready to examine your specimen. During the examination/observation,  
minor features of alignment such as eucentricity and astigmatism should be checked and  
corrected from time to time. Take notes as detail and as much as possible while examining  
specimen and taking photographs.  
Happy Hunting and Enjoying the Small World!  
10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
WAC 09/28/02  
VI. COMPUTER ALIGNMENT AND ASTGMATISM CORRECTION FOR 200 kV  
1. Preses PARAMETERS on TEM bright field page 3 times  
2. Press REMOTE CONTROL to highlight it (This will allow the computer to talk to the TEM).  
3. Turn on the PC monitor (on/off switch on back of monitor).  
4. If computer is already in execution mode, go to step 6  
5. At the prompt, C:\RCM > type CMONITOR  
6. Press F4 to set alignment. When program asks for a file name, choose the file that has a file  
name of the most recent date and type this in (this is the alignment saved by Philips service  
engineer). Answer Y to the question to execute the program and bring you back to the  
cmointor page  
7. Press F6 to set stigmator. When program asks for a file name, choose the file that has the file  
name of the most recent date and type this in. Answer Y to the question to execute the  
program and bring back to the monitor page  
8. Turn off monitor  
9. Press REMOTE CONTROL on third parameters page so that it is not highlighted  
10. Press READY to go back to TEM bright field page.  
11  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Peavey Guitar Nitro II User Manual
Petsafe Electric Pet Fence PIF 300 User Manual
Philips Automobile Alarm 42HF7945D User Manual
Philips MP3 Docking Station AJ300D 79 User Manual
Philips Portable Speaker MCM128B User Manual
Polycom Universal Remote MR16 User Manual
Poulan Lawn Mower Accessory 402337 User Manual
Powermate Portable Generator PM0606750 User Manual
Pride Mobility Mobility Aid 1107 User Manual
PYLE Audio Car Video System PL56PHRB User Manual