Digital Automatic Blood Pressure Monitor
Model i-C10
EN
FR
DE
IT
• Instruction Manual
• Mode d’emploi
• Gebrauchsanweisung
• Manuale di istruzioni
• Manual de instrucciones
• Gebruiksaanwijzing
ES
NL
RU
AR
• ꢀꢁꢂꢃꢄꢃJWXꢄꢃ Uꢃ ꢅꢂWUQꢁFXF\NN
IM-HEM-7070-E-04-05/2012
5320473-7C
Download from Www.Somanuals.com. All Manuals Search And Download.
Before using the unit
Introduction
Thank you for purchasing the OMRON i-C10 Intellisense Upper Arm
Blood Pressure Monitor.
The OMRON i-C10 Intellisense is a fully automatic blood pressure
monitor, operating on the oscillometric principle. It measures your
blood pressure and pulse rate simply and quickly. For comfortable
controlled inflation without the need for pressure pre-setting or re-
inflation the device uses its advanced “Intellisense” technology.
The monitor stores measurement results for two people and features
morning and evening averages. The auto mode allows for three
consecutive measurements with timed intervals for an average
reading.
Please read this instruction manual thoroughly before
using the unit. For specific information about your own
blood pressure, CONSULT YOUR DOCTOR.
EN
3
Download from Www.Somanuals.com. All Manuals Search And Download.
Important Safety Information
Consult your doctor during pregnancy, arrhythmia and arteriosclerosis. Please
read this section carefully before using the unit.
Warning:
Indicates a potentially hazardous situation which, if not avoided, could result in
death or serious injury.
(General Usage)
• Always consult your doctor. Self-diagnosis of measurement results and
selftreatment are dangerous.
• People with severe blood flow problems, or blood disorders, should consult a
doctor before using the unit. Cuff inflation can cause internal bleeding.
(Battery Usage)
• If battery fluid should get in your eyes, immediately rinse with plenty of clean
water. Consult a doctor immediately.
(Optional AC Adapter Usage)
• Never plug in or unplug the power cord from the electric outlet with wet
hands.
Caution:
Indicates a potentially hazardous situation which, if not avoided, may result in
minor or moderate injury to the user or patient or damage to the equipment or
other property.
(General Usage)
• Do not leave the unit unattended with infants or persons who cannot express
their consent.
• Do not use the unit for any purpose other than measuring blood pressure.
• Do not disassemble the unit or arm cuff.
• Do not inflate the arm cuff over 299 mmHg.
• Do not use a mobile phone, or other devices that emit electromagnetic fields,
near the unit. This may result in incorrect operation of the unit.
• Do not operate unit in a moving vehicle (car, airplane).
4
Download from Www.Somanuals.com. All Manuals Search And Download.
Important Safety Information
(Optional AC Adapter Usage)
• Use only the original AC adapter designed for this unit. Use of unsupported
adapters may damage and/or may be hazardous to the unit.
• Plug the AC adapter into the appropriate voltage outlet.
• Do not use the AC adapter if the unit or the power cord is damaged. Turn off
the power and unplug the power cord immediately.
(Battery Usage)
• If battery fluid should get on your skin or clothing, immediately rinse with
plenty of clean water.
• Use only four “AA” alkaline batteries with this unit. Do not use other types of
batteries.
• Do not insert the batteries with their polarities incorrectly aligned.
• Replace old batteries with new ones immediately. Replace all four batteries at
the same time.
• Remove the batteries if the unit will not be used for three months or more.
• When the batteries are replaced, you may need to reset the date and time. If
the year is flashing on the display screen, refer to “2.2 Setting the Date and
Time”.
• Do not use new and used batteries together.
General Safety Precautions
• Do not forcibly bend the arm cuff or bend the air tube excessively.
• When removing the air tube, pull on the air plug at the connection with the
main unit not the tube itself.
EN
• Do not inflate the arm cuff when it is not wrapped around your arm.
• Do not apply strong shocks and vibrations to or drop the unit.
• Do not take measurements after bathing, drinking alcohol, smoking,
exercising or eating.
• Do not wash the arm cuff or immerse it in water.
• Read and follow the “Important information regarding Electro Magnetic
Compatibility (EMC)” in the Technical Data Section.
• Read and follow the “Correct Disposal of This Product” in the Technical Data
Section when disposing of the device and any used accessories or optional
parts.
Save these instructions for future reference.
5
Download from Www.Somanuals.com. All Manuals Search And Download.
1. Overview
Main Unit
I
J
H
G
A
F
L
E
K
D
B
C
A. Display
G. User ID symbol (A or B)
H. Cuff compartment
I. Monitor cover
B. SET button
C. MEMORY button (
)
D. Morning/Evening Average
J. Air jack
buttons (
/
)
K. Battery compartment
L. AC adapter jack (for optional
AC adapter)
E. O/I START button
F. User ID selection switch
(A or B)
6
Download from Www.Somanuals.com. All Manuals Search And Download.
1. Overview
Arm Cuff
M
N
O
M. Arm cuff
(Medium cuff: arm circumference 22-42 cm)
N. Air Plug
O. Air Tube
EN
7
Download from Www.Somanuals.com. All Manuals Search And Download.
1. Overview
Display
P
X
Y
Q
Z
R
T
AA
AC
AE
S
AB
AD
U
V
W
P. Systolic blood pressure
Q. Diastolic blood pressure
R. Heartbeat symbol
X. User ID symbol (A or B)
(Indicates user A or user B as
selected by the user to take a
measurement or when using
the memory function.)
Y. Pulse display
1. (Flashes during measurement)
2.(If flashing after measurement
completed indicates blood
pressure out of recommended
range)
Z. Early morning hypertension
symbol
(Displayed if the morning
average reading for a week is
above 135 for the systolic
blood pressure value and/or 85
for the diastolic blood pressure
value.)
S. Movement symbol
(Displayed if you move your
body during the measurement.)
T. Irregular heartbeat symbol
U. Deflation symbol
V. Battery low symbol
W. Auto mode symbol
(Displayed when Auto Mode is
set.)
AA.Morning average symbol
(Displayed when viewing
morning averages using the
memory function.)
AB.Evening average symbol
(Displayed when viewing
evening averages using the
memory function.)
AC.Average value symbol
(Displayed when viewing value
for the last three
measurements)
AD.Memory symbol
(Displayed when viewing
values stored in memory)
AE.Week display
AF.Date/Time display
8
Download from Www.Somanuals.com. All Manuals Search And Download.
1. Overview
Package Contents
AG
AG.Four “AA” alkaline (LR6)
batteries
• Instruction manual
• Guarantee card
• Blood pressure pass
• Blood Pressure Monitor unit
EN
9
Download from Www.Somanuals.com. All Manuals Search And Download.
2. Preparation
2.1 Installing/Replacing the Batteries
1. Turn the main unit upside down.
2. Slide the battery cover in the
direction of the arrow while
pressing the ribbed part of the
cover.
Battery Cover
3. Install or replace four “AA” size
batteries so that the + (positive)
and - (negative) polarities
match the polarities indicated
on the battery compartment.
4. Put the battery cover back in place.
Slide the battery cover as indicated, until it clicks into place.
Note: The measurement values continue to be stored in memory even
after the batteries are replaced.
10
Download from Www.Somanuals.com. All Manuals Search And Download.
2. Preparation
Battery Life & Replacement
If the battery low symbol ( ) appears on the display, replace all four
batteries at the same time.
- When the battery low symbol ( ) starts to blink, you will still be
able to use the unit for a short while. You should replace the
batteries with new ones ahead of time.
- When the symbol ( ) remains lit, the batteries are exhausted. You
should replace the batteries with new ones at once.Turn the unit off
before replacing the batteries.
• Remove the batteries if the unit will not be used for three months or
more.
• If the batteries have been removed for more than 30 seconds, the
Date/Time setting will need to be reset. See Chapter 2.2 for details.
• Dispose of batteries according to applicable local regulations.
Four new “AA” alkaline batteries will last for approximately
1400 measurements, when used to take six measurements a day.
Since the supplied batteries are for monitoring use only, they may have
a shorter life and not last for 1400 measurements.
EN
11
Download from Www.Somanuals.com. All Manuals Search And Download.
2. Preparation
2.2 Setting the Date and Time
Your blood pressure monitor automatically stores up to 84 individual
measurement values with the date and time.
To make use of the memory and average values functions:
• Set the unit to the correct date and time before taking a measurement
for the first time.
• If the batteries have been removed for more than 30 seconds, the
Date/Time setting will need to be reset.
1. When the O/I START button is
pressed to turn the unit on for the
first time after inserting batteries,
the year digits (2007) will flash on
the display.
2. Press the MEMORY button (
)
to advance the digits one at a time.
Notes:
• The range for the year setting is
2007 to 2030. If the year reaches
2030, it will return to 2007.
• If you hold down the MEMORY
button (
), the digits will
advance rapidly.
3. Press the SET button to confirm
the setting when the desired
number appears on the display.
The year is set and the month
digits flash on the display.
12
Download from Www.Somanuals.com. All Manuals Search And Download.
2. Preparation
4. Repeat steps 2 and 3 to set the
month.
The month is set and the day digits
flash on the display.
5. Repeat steps 2 and 3 to set the
day.
The day is set and the hour digits
flash on the display.
6. Repeat steps 2 and 3 to set the
hour.
The hour is set and the minutes
digits flash on the display.
7. Repeat steps 2 and 3 to set the
minutes.
The minutes settings is set.
EN
The unit automatically turns itself
off after the minute setting has
been set.
To adjust the date and time, press
and hold the SET button while the
unit is turned off.
13
Download from Www.Somanuals.com. All Manuals Search And Download.
2. Preparation
Notes on Adjusting the Date and Time Setting
If you need to adjust the date and time for some reason, or if the date
and time has been reset after replacing the batteries, adjust the date
and time between 10:00 and 18:59. This will avoid problems with
morning and evening weekly averages stored in memory.
To adjust the date or time at a later stage, start the unit in the off
position. Follow the steps below.
1. Press and hold the SET button.
The auto mode symbol
and the current setting (on or
oFF) will appear on the display for the user ID selected.
If the year is not flashing, press the SET button once or twice
until the year is flashing.
The year flashes on the display.
2. Refer to Chapter 2.2.2 to start setting the date and time.
With this manual setting of date and time the unit needs to be
manually shut off. It will not automatically turn off after the
minutes have been set.
14
Download from Www.Somanuals.com. All Manuals Search And Download.
2. Preparation
Notes:
• Please ensure that the date and time have been set correctly so that
the weekly averages will be displayed correctly. If the Date/Time
settings have been accidentally set incorrectly, press the SET button
to circle through the settings in order to adjust to their correct values.
• In the case where the time displays incorrectly after a long period of
non-use of the unit, reset the Date and Time (refer to Chapter 2.2 for
instructions on setting the clock values). Erase the recorded values
when reusing the unit (refer to Chapter 3.5 for directions on erasing
the recorded values).
EN
15
Download from Www.Somanuals.com. All Manuals Search And Download.
Operating instructions
3. Using the Unit
3.1 How to Sit Correctly When Taking a Measurement
You can take a measurement on either your left or right arm.
Notes:
• Measurements should be taken in a quiet place and you should be in
a relaxed, seated position. Make sure that the room is not too hot or
cold.
• Avoid eating, smoking, or exercising for at least 30 minutes before
taking a measurement.
• Do not move or talk during measurement.
Correct
Posture
Remove tight fitting clothing from
your upper arm, and any thick
clothing such as a sweater.
Do not place the cuff over thick
clothes and do not roll up your sleeve
if it is too tight.
Sit upright
with your
back straight.
Place your arm on a table so
that the cuff will be at the same
level as your heart.
The gap between the
chair and the top of the
table should be between
25cm to 30cm.
16
Download from Www.Somanuals.com. All Manuals Search And Download.
3. Using the Unit
Notes:
• Correct posture during measurement is necessary to get accurate
results.
• You should also try to measure your blood pressure at the same time
each day. (Within 1 hour after waking up is recommended.)
Incorrect Posture
• Arched back (leaning
forwards)
• Sitting cross-legged
• Sitting on a sofa or at a low
table so that you tend to lean
forward
These situations could lead to higher blood pressure values due to
strain or the arm cuff being lower than the heart.
If the arm cuff is at a lower position than your heart use cushions
etc., to adjust the height of your arm.
EN
17
Download from Www.Somanuals.com. All Manuals Search And Download.
3. Using the Unit
3.2 Applying the Arm Cuff
You can wrap the cuff either on your right or left arm.
• Remove tight-fitting clothing from your
upper arm.
• Do not place the cuff over thick clothes and
do not roll up your sleeve if it is too tight.
Taking Measurements on the Left Arm
1. Insert the air plug into the air jack (inside the Cuff Compartment
of the device).
Air plug
Air jack
The cuff must be fully deflated when it is inserted into the air
jack.
2. Sit on a chair with your feet flat on the floor and place your arm
on a table so that the cuff will be at the same level as your
heart.
18
Download from Www.Somanuals.com. All Manuals Search And Download.
3. Using the Unit
3. Hold the grip on the cuff securely
Grip
with your hand.
Fabric fastener
4. Turn the palm of your hand upward.
5. Apply the cuff to your upper arm so that the blue strip is centred
on the middle of your inner arm and points down the inside of
the arm. The air tube should run down the inside of your
forearm and be in line with your middle finger.
The bottom of the cuff should be approximately 1 to 2 cm
above the elbow.
Bottom edge 1 to 2 cm
Blue strip
EN
Air tube
19
Download from Www.Somanuals.com. All Manuals Search And Download.
3. Using the Unit
6. When the cuff is positioned correctly,
close the fabric fastener FIRMLY.
Notes:
• Make certain the cuff fits snugly around your arm.
• The cuff should make good contact with your skin. You should
be able to fit your index finger between the cuff and your arm
easily, so you can pull the cuff off and on.
• Make sure that there are no kinks in the air tubing.
Taking Measurements on the Right Arm
Please note the following points when
applying the cuff to your right arm. Apply the
cuff so that the air tube is at the side of your
elbow.
Notes:
• Be careful not to rest your arm on the air
tube, or otherwise restrict the flow of air to
the cuff.
• Apply the arm cuff so that no part of the
cuff is positioned over the elbow joint. The
cuff should be 1 to 2 cm above the elbow.
• If used 2 times per day, the cuff will last for 3 years.
• The blood pressure can differ between the right arm and the left arm,
and therefore also the measured blood pressure values can be
different. Omron recommends to always use the same arm for
measurement. If the values between the two arms differ substantially,
please check with your physician which arm to use for your
measurement.
20
Download from Www.Somanuals.com. All Manuals Search And Download.
3. Using the Unit
3.3 Taking a Reading
The unit is designed to take measurements and store the
measurement values in the memory for two people using user ID A and
user ID B.
The unit can also be used to take a single measurement for other
persons using the guest mode. The user ID symbol is not displayed
when using the guest mode. Measurements taken using the guest
mode are not stored in the memory.
Using the Guest Mode
1. Press and hold the O/I START button until the A and B user
symbols disappear from the display.
A and B display
goes out
Press for at least 3 seconds
EN
2. Release the O/I START button.
The cuff starts to inflate automatically.
Note: To stop the inflation press and release the O/I START
button. The unit will stop inflating, start deflation and turn
off.
3. Inflation stops and the measurement is started. When the
measurement is completed, your blood pressure and pulse rate
appear on the display.
21
Download from Www.Somanuals.com. All Manuals Search And Download.
3. Using the Unit
Selecting the User ID
Select your user ID. Always use the same user ID when taking a
measurement. The unit stores the measurement values in the memory
by the user ID selected. These measurement values are used to
calculate the morning averages and evening averages.
Slide the user ID selection switch up or down to select user A or B.
Using the User ID
When you select a user ID the unit can be set to take one
measurement using the single mode or three consecutive
measurements using the auto mode when you press the O/I START
button.
Note: The default value is set to the single mode to take one
measurement.
The auto mode takes three consecutive measurements. The unit will
inflate the cuff and take three measurements separated by a short
interval of time between each measurement.
22
Download from Www.Somanuals.com. All Manuals Search And Download.
3. Using the Unit
Using the Single Mode
1. Select your user ID (A or B).
2. Press the O/I START button.
All symbols appear on the display. The user ID symbol you
selected flashes on the display.
The cuff starts to inflate automatically. As the cuff inflates, the
unit automatically determines your ideal inflation level. This unit
detects the pulse during inflation. Do not move your arm and
remain still until the entire measurement process is completed.
EN
Note: To stop the inflation or measurement, press and release
the O/I START button. The unit will stop inflating, start
deflating and will turn off.
23
Download from Www.Somanuals.com. All Manuals Search And Download.
3. Using the Unit
3. Inflation stops automatically and the measurement is started.
As the cuff deflates, decreasing numbers appear on the
display. The heartbeat symbol ( ) flashes at every heartbeat.
4. When the measurement is complete, the arm
cuff completely deflates. Your blood pressure
and pulse rate are displayed.
5. Press the O/I START button to turn
the unit off.
Note: If you forget to turn the
monitor off, it will
automatically turn off after
five minutes.
Notes:
• Self-diagnosis of measured results and treatment are
dangerous. Please follow the instructions of your doctor.
• Wait 2-3 minutes before taking another blood pressure
measurement. Waiting between readings allows the arteries
to return to the condition prior to taking the blood pressure
measurement.
24
Download from Www.Somanuals.com. All Manuals Search And Download.
3. Using the Unit
Selecting the Auto Mode
Auto Mode may be selected for either A or B user.
1. Make sure the unit is powered off.
2. Select your user ID (A or B).
3. Press and hold the SET button.
The auto mode symbol (
the display.
) and the oFF setting appears on
EN
4. Press the MEMORY button
(
) to select on.
Note: Press the MEMORY
button (
) to alternate
between on and oFF.
25
Download from Www.Somanuals.com. All Manuals Search And Download.
3. Using the Unit
5. Press the SET button to confirm the change.
The number of seconds between each measurement is
displayed.
Note: The default value between each measurement is
60 seconds.
6. Press the MEMORY button (
) to change the interval of
time between each measurement.
The interval can be set to 15 seconds, 30 seconds, 60 seconds
or 120 seconds. Continue to press the MEMORY button (
until the desired interval appears on the display.
)
7. Press the SET button to confirm the
setting.
The year flashes on the display.
8. Press the O/I START button to turn
the unit off.
26
Download from Www.Somanuals.com. All Manuals Search And Download.
3. Using the Unit
Using the Auto Mode
1. Select your user ID (A or B).
2. Press the O/I START button.
All symbols appear on the display. The user ID Symbol you
selected flashes on the display. The cuff starts to inflate
automatically.
Note: To stop the inflation or measurement, press the O/I
START button. The unit will stop inflating, start deflating
and will turn off.
3. Inflation stops automatically and the measurement is started.
As the cuff deflates, decreasing numbers appear on the
display. The heartbeat symbol ( ) flashes at every heartbeat.
When the measurement is complete, the arm cuff completely
deflates.
Note: Do not move your arm and remain still until the entire
measurement process is completed.
EN
4. Check the progress of the consecutive measurements.
The first measurement is complete, waiting to start the next
measurement. The interval of time set between each
measurement will appear on the display.
27
Download from Www.Somanuals.com. All Manuals Search And Download.
3. Using the Unit
The second measurement starts automatically.
The second measurement is complete, waiting to start the next
measurement.
The third measurement starts automatically.
The consecutive measurements are complete. The average for
the three measurements appears on the display.
The three individual measurement results are not displayed
while Auto Mode is taking the measurements. However, the
results for each individual measurement can be displayed after
all the measurements have been completed.
Note: During the Auto Mode the unit will repeat the
measurement in case of an error. After 3 errors the unit
will stop taking further measurements. If only two
measurements are successful the average of these two
measurements will be displayed. If only one
measurement is successful, this result will be displayed.
5. Press the SET button to display the blood pressure and pulse
rate for the first measurement. Press the SET button again to
display the next measurement.
6. Press the O/I START button to turn the unit off.
28
Download from Www.Somanuals.com. All Manuals Search And Download.
3. Using the Unit
Important:
• If your systolic or diastolic pressure is outside the
standard range, the heartbeat symbol will blink when
the measurement result is displayed.
Recent research suggests that the following values
can be used as a guide to high blood pressure for
measurements taken at home.
Systolic Blood Pressure
Above 135 mmHg
Diastolic Blood Pressure
Above 85 mmHg
This criteria is for home blood pressure measurement.
For professional office blood pressure measurement criteria, please
refer to Chapter 9.
• Your blood pressure monitor includes an irregular
heartbeat feature. Irregular heartbeats can influence
the results of the measurement. The irregular
heartbeat algorithm automatically determines if the
measurement is usable or needs to be repeated. If
the measurement results are affected by irregular
heartbeats but the result is valid, the result is shown
together with the irregular heartbeat icon. If the irregular heartbeats
cause the measurement to be invalid, no result is shown. If the
irregular heartbeat symbol (
) is shown after you have taken a
EN
measurement, repeat the measurement. If the irregular heartbeat
icon is shown frequently, please make your doctor aware of it.
29
Download from Www.Somanuals.com. All Manuals Search And Download.
3. Using the Unit
What is Irregular Heartbeat?
Normal Heartbeat
Irregular Heartbeat
An irregular heartbeat is a heartbeat
rhythm that varies by more than 25%
from the average heartbeat rhythm
detected while the unit is measuring
the systolic and diastolic blood
pressure.
Pulse
Blood pressure
Long
Short
Pulse
If such an irregular rhythm is
detected more than twice during
measurement, the irregular
Blood pressure
heartbeat symbol (
) appears on
the symbol when the measurement
results are displayed.
What is Arrhythmia?
A heartbeat is stimulated by electrical signals that cause the heart to
contract.
Arrhythmia is a condition where the heartbeat rhythm is abnormal due
to flaws in the bio-electrical system that drives the heartbeat. Typical
symptoms are skipped heartbeats, premature contraction, an
abnormally rapid (tachycardia) or slow (bradycardia) pulse. This can
be caused by heart disease, aging, physical predisposition, stress,
lack of sleep, fatigue etc. Arrhythmia can only be diagnosed by a
doctor through a special examination.
Whether the appearance of the irregular heartbeat symbol (
) in
the results indicates arrhythmia or not can only be determined by an
examination and diagnosis by your doctor.
If the irregular heartbeat symbol (
) is shown frequently, please
make your doctor aware of it. Conducting self-diagnosis and
treatment based on measurement results is dangerous. Be sure to
follow the instructions of your doctor.
30
Download from Www.Somanuals.com. All Manuals Search And Download.
3. Using the Unit
3.4 Instructions for Special Conditions
If your systolic pressure is known to be more than 220 mmHg, press
and hold the O/I START button until the arm cuff inflates 30 to
40 mmHg higher than your suspected systolic pressure.
1. Press the O/I START button to turn
the unit on.
Measurement starts.
2. When the cuff starts to inflate, press the O/I START button and
keep it pressed until the pressure is 30 to 40 mmHg higher
than your suspected systolic pressure.
Note: You cannot inflate the cuff above 299 mmHg. (An error
will be displayed if you try to inflate the cuff above
300 mmHg.)
3. Release the O/I START button when the cuff has been inflated
to the desired pressure.
EN
The cuff starts to deflate and measurement starts.
4. The rest of the procedure is the same as for normal
measurement. Refer to Chapter 3.3.
Note: Do not apply more pressure than necessary.
31
Download from Www.Somanuals.com. All Manuals Search And Download.
3. Using the Unit
3.5 Using the Memory Function
The unit is designed to store the blood pressure and the pulse rate in
the memory for two people (user A and user B) every time a
measurement is completed.
The unit automatically stores up to 84 sets of measurement values
(blood pressure and pulse rate) for each user (A and B). When 84 sets
of measurement values are stored, the oldest record is deleted to save
the most recent values. The unit also stores 8 weeks of morning
averages and 8 weeks of evening averages for each user (A and B).
Average Function
The unit calculates the average reading based on the three most
recent sets of measurement values taken within 10 minutes of the most
recent reading.
Note: If two sets of measurement values are stored in the memory for
the 10 minutes period, the average is based on the two sets of
measurement values. If one set of measurement values is
stored, this is displayed as the average.
To Display the Measurement Values
1. Select your user ID (A or B).
2. Press the MEMORY button (
).
The average value is shown and the average symbol ( )
appears above the memory symbol ( ) on the display.
32
Download from Www.Somanuals.com. All Manuals Search And Download.
3. Using the Unit
3. Press the MEMORY button (
) to display the most recent
set of individual measurement values on the screen. The
values are displayed from the most recent to the oldest.
Note: The date and time are
alternately displayed with the
measurement values.
Alternating display
Press the MEMORY button (
values.
) to display the next set of
Press and hold the MEMORY button (
values faster.
) to display the
Auto mode
The average for the consecutive measurements appears on
the display with the auto mode symbol (
). To display the
values for the individual measurements press the SET button
while the average is displayed.
EN
4. Press the O/I START button to turn the unit off.
33
Download from Www.Somanuals.com. All Manuals Search And Download.
3. Using the Unit
Morning and Evening Averages
The unit calculates and displays a weekly average for measurements
taken in the morning ( ) and evening ( ). The unit stores 8 weeks of
morning averages and 8 weeks of evening averages for each user (A
and B).
Note: The week begins Sunday at 4:00.
Morning Averages
Morning averages are based on the first average reading using the
auto mode or the first reading using the single mode.
Measurement times must be between:
4:00 and 9:59
Evening Averages
Evening averages are based on the last average reading using the
auto mode or the last reading using the single mode.
Measurement times must be between:
19:00 and 1:59
34
Download from Www.Somanuals.com. All Manuals Search And Download.
3. Using the Unit
About the Weekly Averages
Morning Weekly Average
This is the average for measurements taken during the morning
(4:00-9:59) between Sunday and Saturday. An average for each day is
calculated for up to three measurements taken within 10 minutes of the
first measurement of the morning.
Measurements of the morning
Sunday
Monday
Tuesday Wednesday Thursday
Friday
Saturday
4:00
First measurement of
the morning
Measurement
Measurement
Measurement
Measurement
Up to 3 measurements
Within 10 minutes
Morning Blood Pressure
Measurement
Measurement
9:59
Evening Weekly Average
This is the average for measurements taken during the evening
(19:00-1:59) between Sunday and Saturday. An average for each day
is calculated for up to three measurements taken within 10 minutes of
the last measurement of the evening.
EN
Measurements of the evening
Sunday
Monday
Tuesday Wednesday Thursday
Friday
Saturday
19:00
Measurement
Measurement
Within 10 minutes
Measurement
Measurement
Measurement
Measurement
Up to 3 measurements
Last measurement of
the evening
Evening Blood Pressure
1:59
35
Download from Www.Somanuals.com. All Manuals Search And Download.
3. Using the Unit
To Display Morning and Evening Averages
1. Select your user ID (A or B).
2. Press the morning average
button (
) or the evening
average button (
).
The average for the current
week “THIS WEEK” appears on
the display.
The morning and evening
average value of the same week
can be displayed by pressing the
morning average button (
)
or the evening average button
(
). It is important to check
both morning and evening
average of the same week.
3. Continue to press the morning average button (
) or the
evening average button (
) to display the previous weeks.
The unit displays “-1 WEEK” for the previous week to
“-7 WEEK” for the oldest set of averages.
Note: If there are not enough measurements
taken in a week to calculate the average
the display will appear as illustrated.
4. Press the O/I START button to turn the unit off.
36
Download from Www.Somanuals.com. All Manuals Search And Download.
3. Using the Unit
Notes on Display Combinations
In addition to the morning average and evening average symbols, the
unit may also display the morning hypertension symbol if the morning
average for that week is above the home blood pressure guideline.
(See Chapter 3 “Using the Unit - Important information” for details.)
Depending on your measurement results, these may be displayed in
the following combinations.
Display
Morning
Average
Morning average from This Week
with Morning Average Symbol
Morning average from 7 weeks
ago with Morning Average
Symbol + blinking heartbeat
Symbol + Morning Hypertension
Symbol
EN
Evening
average
Evening average from This Week Evening average from 7 weeks
with Evening Average Symbol
ago with Evening Average
Symbol + blinking Heartbeat
Symbol + Morning Hypertension
Symbol
Average within home blood
pressure guideline
Average above home blood
pressure guideline + Morning
Hypertension
37
Download from Www.Somanuals.com. All Manuals Search And Download.
3. Using the Unit
The morning hypertension symbol ( ) appears if the weekly average
for morning measurements is above 135/85.
In this case the morning hypertension symbol ( ) is displayed when
the evening average is displayed, regardless of the values for the
evening average.
To Delete All Values Stored in the Memory
The values stored in the memory are deleted by user ID.
You cannot partially delete values stored in the memory. All values for
the user you select will be deleted.
1. Select your ID (A or B).
2. While holding the MEMORY button (
) press the O/I START
button simultaneously for more than 2 seconds and all values
will be deleted.
38
Download from Www.Somanuals.com. All Manuals Search And Download.
4. Quick Reference Guide
This Quick Reference Guide is for reference only. Please read
the entire manual before using the unit for the first time.
To help ensure a reliable reading, avoid eating, smoking, or
exercising for at least 30 minutes before taking a measurement.
Note: Remove any tight-fitting clothing from your upper arm.
1. Sit on a chair with your feet flat on the floor and place your arm
on a table so that the arm cuff will be at the same level as your
heart.
2. Secure the cuff around your arm using the fabric fastener strip.
3. Select your user ID (A or B).
EN
4. Press the O/I START button.
When measurement is complete, the
unit displays your blood pressure and
pulse rate, and automatically deflates
the cuff.
Notes:
• Always wait at least 2-3 minutes before taking another blood
pressure measurement.
• Please note that all measurement results are stored in the
memory. If different people take measurements using the
same device, make sure that you are aware of this fact.
39
Care and maintenance
Download from Www.Somanuals.com. All Manuals Search And Download.
5. Handling Errors and Problems
5.1 Error Messages
Error Display
Cause
Remedy
Cuff is under inflated.
Movement during
measurement.
Carefully read and repeat the steps
listed under Chapter 3.3.
Cuff over inflated.
Replace all four “AA” batteries with
new ones.
Refer to Chapter 2.1.
Batteries have run down.
Blinks or appears
continuously.
Carefully read and repeat the steps
listed under Chapter 3.3.
Movement, sound.
Cuff leaks air.
Replace with a new cuff.
40
Download from Www.Somanuals.com. All Manuals Search And Download.
5. Handling Errors and Problems
5.2 Troubleshooting
Problem
Cause
Remedy
No display appears
when the O/I START
button is pressed.
The batteries are empty.
Is the cuff wrapped
Replace the batteries.
Attach the arm cuff
correctly around the arm? correctly.
Cannot measure or
The cuff pressure does
not rise although the
pump motor can be
heard.
Check that the air plug is
readings are too high.
properly connected to the
unit. Push the air plug
firmly into the air jack.
The blood pressure is
different each time. The
reading is extremely low
(or high).
Blood pressure readings constantly vary with time of
measurement and nervous condition. Take deep
breaths to relax before taking a measurement.
EN
41
Download from Www.Somanuals.com. All Manuals Search And Download.
6. Maintenance and Storage
To protect your unit from damage, please observe the following:
• Do not subject your unit and cuff to extreme temperatures, humidity,
moisture or direct sunlight.
• Do not fold the cuff or tubing tightly.
• Do not disassemble the unit.
• Do not carry out repairs of any kind yourself. If a defect occurs,
consult the OMRON distributor or Customer Services as mentioned
on the packaging, or seek the advice of your surgical appliance
stockist or pharmacist.
• The OMRON i-C10 is a precision measuring device. It is important
that the accuracy remains within the specifications. It is recommended
that you have the unit inspected every two years to ensure it is
accurate and functioning correctly. Consult the OMRON distributor or
Customer Services as mentioned on the packaging for further details.
• Do not subject the unit to strong shocks or vibrations (for example,
dropping the unit on the floor.)
• Do not use volatile liquids to clean the unit. THE UNIT SHOULD BE
CLEANED WITH A SOFT, DRY CLOTH.
• Do not wash the arm cuff or immerse it in water.
• Use a soft, moistened cloth and soap to clean the arm cuff.
Storage
Close the monitor cover when not in use.
42
Download from Www.Somanuals.com. All Manuals Search And Download.
7. Optional Parts
AC Adapter "R"
AC Adapter "Q"
Arm Cuff
R Adapter-9997605-0
Q Adapter-1098336-8
Order No. 9956685-4
Using the Optional AC Adapter
Insert batteries into the battery compartment even when using the AC
adapter.
Notes:
• Never plug in or unplug the power cord from the electric outlet with
wet hands.
• Use only the original AC adapter designed for this unit. Use of
unsupported adapters may damage the unit.
• When storing the AC adapter with the main unit, be careful not to
damage the main unit or cuff.
1. Insert the AC adapter plug into
the AC adapter jack on the left
side of the main unit.
EN
2. Plug the AC adapter into an
electrical outlet.
To disconnect the AC adapter, unplug the AC adapter from the
electrical outlet first and then immediately remove the AC
adapter plug from the main unit.
43
Download from Www.Somanuals.com. All Manuals Search And Download.
8. Technical Data
Product Description
Model
Digital Automatic Blood Pressure Monitor
OMRON i-C10 (HEM-7070-E)
LCD Digital Display
Display
Measurement Method
Measurement Range
Oscillometric method
Pressure: 0 mmHg to 299 mmHg
Pulse: 40 to 180/min.
Accuracy
Pressure: ꢀ3 mmHg
Pulse: ꢀ 5% of display reading
Inflation
Deflation
Pressure Detection
Memory
Fuzzy-logic controlled by electric pump
Automatic pressure release valve
Capacitive pressure sensor
84 Measurements with date and time for each user
(A and B)
Power Source
Battery Life
4 “AA” batteries 1.5V or AC/ DC adapter (6V = 4W)
Capacity of new alkaline batteries is approx. 1400
measurements when used 6 times a day with 4 new
alkaline batteries
Operating temperature/
Humidity
+10°C to +40°C
Maximum: 30 to 90% RH
Storage temperature/
Humidity/Air Pressure
-20°C to +60°C
Maximum: 10 to 95% RH
700 - 1060 hPa
Console Weight
Cuff Weight
Approximately 960g without batteries
Approximately 240g
Outer Dimensions
Cuff Dimensions
Approximately 249 (l) mm ꢁ 178 (w) mm ꢁ 110 (h) mm
Approximately 152 mm ꢁ 600 mm
(Cuff: arm circumference 22 to 42 cm)
Package Content
Cuff, instruction manual, battery set, guarantee card, blood
pressure pass
Note: Subject to technical modification without prior notice.
= Type B
= Type B
• This device fulfils the previsions of EC directive 93/42/EEC (Medical Device Directive).
• This blood pressure monitor is designed according to the European Standard EN1060,
Non-invasive sphygmomanometers Part 1: General Requirements and Part 3:
Supplementary requirements for electromechanical blood pressure measuring
systems.
• This OMRON product is produced under the strict quality system of OMRON
Healthcare Co. Ltd., Japan. The Core component for OMRON blood pressure
monitors, which is the Pressure Sensor, is produced in Japan for assembly.
44
Download from Www.Somanuals.com. All Manuals Search And Download.
8. Technical Data
Important information regarding Electro Magnetic Compatibility (EMC)
With the increased number of electronic devices such as PC’s and mobile (cellular)
telephones, medical devices in use may be susceptible to electromagnetic
interference from other devices. Electromagnetic interference may result in incorrect
operation of the medical device and create a potentially unsafe situation.
Medical devices should also not interfere with other devices.
In order to regulate the requirements for EMC (Electro Magnetic Compatibility) with
the aim to prevent unsafe product situations, the EN60601-1-2 standard has been
implemented. This standard defines the levels of immunity to electromagnetic
interferences as well as maximum levels of electromagnetic emissions for medical
devices.
This medical device manufactured by OMRON Healthcare conforms to this
EN60601-1-2:2001 standard for both immunity and emissions.
Nevertheless, special precautions need to be observed:
• Do not use mobile (cellular) telephones and other devices, which generate strong
electrical or electromagnetic fields, near the medical device. This may result in
incorrect operation of the unit and create a potentially unsafe situation.
Recommendation is to keep a minimum distance of 7 m. Verify correct operation of
the device in case the distance is shorter.
Further documentation in accordance with EN60601-1-2:2001 is available at
OMRON Healthcare Europe at the address mentioned in this instruction manual.
Correct Disposal of This Product
(Waste Electrical & Electronic Equipment)
This marking shown on the product or its literature, indicates that it
should not be disposed of, with other household wastes at the end of
its working life. To prevent possible harm to the environment or human
health from uncontrolled waste disposal, please separate this from
other types of wastes and recycle it responsibly to promote the
sustainable reuse of material resources.
EN
Household users should contact either the retailer where they purchased this product,
or their local government office, for details of where and how they can take this item
for environmentally safe recycling.
Business users should contact their supplier and check the terms and conditions of
the purchase contract. This product should not be mixed with other commercial
wastes for disposal.
This product does not contain any hazardous substances.
45
Download from Www.Somanuals.com. All Manuals Search And Download.
9. Some Useful Information about Blood Pressure
What is Blood Pressure?
Blood pressure is a measure of the force of blood flowing against the
walls of the arteries. Arterial blood pressure is constantly changing
during the course of the heart's cycle.
The highest pressure in the cycle is called the Systolic Blood Pressure;
the lowest is the Diastolic Blood Pressure.
Both pressure readings, the Systolic and Diastolic, are necessary to
enable a doctor to evaluate the status of a patient's blood pressure.
Why is it a Good Thing to Measure Blood Pressure at Home?
Having your blood pressure measured by a doctor can cause anxiety
which is itself a cause of high blood pressure. As a variety of conditions
affect blood pressure, a single measurement may not be sufficient for
an accurate diagnosis.
Many factors such as physical activity, anxiety, or the time of day, can
influence your blood pressure. Thus it is best to try and measure your
blood pressure at the same time each day, to get an accurate
indication of any changes in blood pressure. Blood pressure is typically
low in the morning and increases from afternoon to evening. It is lower
in the summer and higher in the winter.
Blood pressure is measured in millimetres of mercury (mmHg) and
measurements are written with the systolic pressure before the
diastolic e.g. A blood pressure written as 140/90, is referred to as
140 over 90 mmHg.
46
Download from Www.Somanuals.com. All Manuals Search And Download.
9. Some Useful Information about Blood Pressure
Upper curve: systolic blood pressure
Lower curve: diastolic blood pressure
200
150
100
50
6
12
18
24
Time of day
Example: fluctuation within a day (male, 35 years old)
Classification of Blood Pressure by the World Health
Organization
The World Health Organization (WHO) and the International Society of
Hypertension (ISH) developed the Blood Pressure Classification
shown in this figure.
Systolic blood pressure
(mmHg)
Severe hypertension
180
Moderate hypertension
160
Mild hypertension
140
EN
Normal systolic value
130
Normal blood
pressure
120
Optimal blood
pressure
(target value)
Diastolic
blood
pressure
(mmHg)
80
85
90
100 110
This classification is based on the blood pressure values measured on
people in a sitting position in outpatient departments of hospitals.
Note: There is no universally accepted definition of hypotension.
However, those having the systolic pressure below 100 mmHg
are assumed as hypotensive.
47
Download from Www.Somanuals.com. All Manuals Search And Download.
9. Some Useful Information about Blood Pressure
Manufacturer
OMRON HEALTHCARE Co., Ltd.
53, Kunotsubo, Terado-cho, Muko, Kyoto,
617-0002 JAPAN
EU-representative
OMRON HEALTHCARE EUROPE B.V.
Scorpius 33, 2132 LR Hoofddorp, THE NETHERLANDS
OMRON (DALIAN) CO., LTD.
Production facility
Subsidiary
Dalian, CHINA
OMRON HEALTHCARE UK LTD.