Professional Blood Pressure Monitor
HBP-1300
• Instruction Manual
• Mode d’emploi
• Gebrauchsanweisung
• Manuale di istruzioni
• Manual de instrucciones
• Gebruiksaanwijzing
EN
FR
DE
IT
ES
NL
RU
TR
AR
•
VYꢂTHTJWXHT ꢀT ꢁꢂWꢀQYFXF\NN
• Kullanꢃm Kꢃlavuzu
Thank you for purchasing this OMRON Professional Blood Pressure Monitor.
Please completely read this Instruction Manual before using the monitor for the first time.
Read this manual to ensure the safe and accurate use of the monitor.
IM-HBP-1300-E-01-02/2013
5337346-6A
Download from Www.Somanuals.com. All Manuals Search And Download.
Introduction
Intended Use
Medical Purpose
The device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult and
pediatric patient population with arm circumference ranging from 12 cm to 50 cm (from 5 inches to
20 inches).
Intended User
This device should be used by a medical professional.
Patient Population
This device is intended for use on adults and children of age 3 years and older.
Environment
The instrument is designed for use in physicians’ offices, hospitals, clinics and other medical facilities.
Measurement Parameter
ꢀ Non-Invasive Blood Pressure
ꢀ Pulse rate
Precautions for Use
Warnings and cautions described in the instruction manual should be observed at all times.
Exemptions
OMRON will not bear any responsibilities on the following matters.
1. When a problem or damage occurs caused by the maintenance and/or repair
conducted by a person other than OMRON or the dealer specified by OMRON
2. The problem or damage of OMRON product caused by the product of other
manufacturer not delivered by OMRON
3. The problem and damage caused by the maintenance and/or repair using the repair
parts not specified by OMRON
4. The problem and damage caused by the results not observing the Notes on Safety or
the operational method mentioned in this Instruction Manual
EN
5. Under the circumstances not within the operating conditions of this unit including the
power source or the setting environment mentioned in this Instruction Manual
6. The problem and damage caused by the result(s) of remodeling or improper repair of
this product
7. The problem and damage caused by act of god such as fire, earthquake, flood, or
lightning
1. The contents of this Instruction Manual may be changed without prior notice.
2. We have thoroughly reviewed the contents of this Instruction Manual.
However, if an inadequate description or error is found, please let us know.
3. It is prohibited to copy a part of or the entire Instruction Manual without getting
OMRON’s permission. Unless this Instruction Manual is used by an individual
(company), it cannot be used without getting OMRON’s permission from the standpoint
of the Copyright Law.
1
Download from Www.Somanuals.com. All Manuals Search And Download.
Notes on Safety
The warning signs and symbol examples indicated below are intended to ensure safe use of the
product and prevent damage and injury to you and others. The signs and symbols are explained below.
Safety Symbols used in this Instruction Manual
Indicates the matters in which death or severe bodily damage may arise as a result
of incorrect handling.
Warning
Indicates the matters in which bodily harm or material damage may arise as a result
of incorrect handling.
Caution
General Information
Indicates general information to keep in mind when using the unit and other useful information.
Note:
2
Download from Www.Somanuals.com. All Manuals Search And Download.
Warnings and Cautions
Usage warnings and cautions
Setup
Warning
Do not use the cuff or AC adapter to lift the unit, it can also cause the unit to malfunction.
If the unit has broken down, contact OMRON HEALTHCARE.
Do not use in combination with a hyperbaric oxygen therapy device, or in an environment where
combustible gas may be generated.
Do not use in combination with magnetic resonance imaging (MRI) equipment. If MRI is to be
performed, remove cuff connected to the unit from the patient.
Do not use with a defibrillator.
Do not install the unit in the following locations:
- Locations subject to vibration such as ambulances and emergency helicopters.
- A location where there is gas or flame.
- A location where there is water or steam.
- A location where chemicals are stored.
Do not use at extremely high temperature, high humidity, or high altitude. Use only within the required
ambient conditions.
Do not subject the unit to intense shock.
Do not place heavy objects on the AC adapter cable, or allow the unit to sit on the cord.
Clinical testing has not been conducted on newborn infants and pregnant women. Do not use on
newborn infants and pregnant women.
Do not plug in or unplug the AC adapter with wet hands.
Caution
EN
Do not install the unit in the following locations:
- Locations with dust, salt, or sulfur.
- Locations directly exposed to sunlight for extended periods of time (in particular, do not leave in
direct sunlight or near a source of ultraviolet light for extended periods, as ultraviolet light will cause
deterioration of the LCD).
- Locations subject to vibration or shock.
- Near heaters.
Do not use the unit near large equipment that uses a switching relay for power ON/OFF.
3
Download from Www.Somanuals.com. All Manuals Search And Download.
Before use / during use
Warning
The unit complies with the EMC standard (IEC60601-1-2). As such, it can be used simultaneously
with multiple medical instruments. However, if instruments that generate noise such as an electric
scalpel or a microwave therapy device are near the unit, check the operation of the unit during and
after use of these instruments.
If an error occurs or a measurement result is questionable, check the vital signs of the patient by
auscultation or palpation. Avoid relying solely on the measurement results of the unit when judging
the patient’s condition.
Only trained healthcare providers should use this device. Do not allow patients to operate this
device.
Properly connect the connectors and AC adapter cable.
Do not place objects or liquids on top of this unit.
Check the following before using the unit:
- Make sure the AC adapter cable is not damaged (wires are not exposed or broken), and the
connections are firm.
For the AC adapter connected to the unit, supplies, and optional devices, use only the standard
accessories or OMRON-specified products.
Do not use in a location with moisture, or a location where water may splash on the unit.
This unit is intended for use in physicians’ office, hospitals, clinics and other medical facilities.
Do not use the unit if it emits smoke, an abnormal odor, or abnormal noise.
Do not bring cellular telephones or transceivers into the room where the unit is installed or being
used.
Do not connect multiple monitors to the same patient.
Do not connect the unit to a power outlet that is controlled by a wall switch.
4
Download from Www.Somanuals.com. All Manuals Search And Download.
Caution
Before using the unit, verify that none of the following apply to the patient:
- Poor peripheral circulation, noticeably low blood pressure, or low body temperature (there will be
low blood flow to the measurement position)
- The patient uses an artificial heart and lung (there will be no pulse)
- An SpO sensor and the cuff are attached to the same arm
2
- The patient has an aneurysm
- The patient has frequent arrhythmia
- Body motions such as convulsions, arterial pulsations, or trembling (cardiac massage in progress,
minute continuous vibrations, rheumatism, etc.)
Before use, visually inspect the unit to make sure there are no deformations due to falling, and that
there is no dirt or moisture on the unit.
When the unit has not been used for an extended period of time, always verify that it operates
normally and safely before use.
Do not use in a location where the unit may easily fall. In the event that the unit falls, verify that it
operates normally and safely.
EN
5
Download from Www.Somanuals.com. All Manuals Search And Download.
Cleaning
Warning
When cleaning the unit, turn off the power and disconnect the AC adapter from the unit.
After cleaning the unit, make sure it is completely dry before connecting to a power outlet.
Do not spray, pour, or spill liquids into or onto the unit, accessories, connectors, buttons, or openings
in the housing.
Caution
Do not use thinner, benzene, or other solvents to clean the unit.
Do not sterilize by autoclave or gas sterilization (EOG, formaldehyde gas, high-concentration ozone,
etc.).
If using an antiseptic solution for cleaning, follow the instructions of the manufacturer.
Clean the unit regularly.
Maintenance and inspection
Warning
To use the unity safely and correctly, always inspect the unit when starting work.
Unauthorized modification is prohibited by law. Do not attempt to disassemble or modify the unit.
6
Download from Www.Somanuals.com. All Manuals Search And Download.
Warnings and cautions for safe measurement
Rechargeable battery
Warning
If battery fluid comes in contact with the eye, immediately flush with copious amounts of water. Do not
rub. Seek medical attention immediately.
Do not use the battery pack in any other device besides this unit. Do not throw into flame,
disassemble, or heat.
Always disconnect the AC adapter from the unit before removing or installing a battery.
If the unit will not be used for a month or longer, remove the battery from the unit and store. Charge
the battery once every 6 months. (Storage conditions for the battery are a temperature of -20 to 30°C
(-4 to 86°F) and a humidity of 65 20%.)
Before use, always charge the battery.
Do not attempt to disassemble or modify the battery.
Do not apply pressure to and deform the battery. Do not throw, pound, drop, bend, or hit the battery.
The battery has positive/negative polarity. If the battery does not connect well to the unit, do not
forcibly connect it.
Do not connect the positive and negative terminals of the battery with a wire or other metal object. Do
not carry or store a battery with metal necklaces, hairpins, or other metal objects.
Use only the specified type of battery.
Caution
Do not touch the positive and negative terminals of the battery pack with a wire or other metal object.
If battery fluid comes into contact with the skin or clothes, immediately rinse with water.
EN
7
Download from Www.Somanuals.com. All Manuals Search And Download.
Non-Invasive Blood Pressure (NIBP) measurement
Warning
If a cuff is used on a patient with an infection, treat the cuff as medical waste, or disinfect before
reuse.
If frequently performing NIBP measurement using a cuff over an extended period of time, periodically
check the patient’s circulation. In addition, wrap the cuff as indicated in the cautionary points in this
manual.
Do not connect the NIBP cuff or cuff joint to a luer lock adapter.
Do not bend cuff tube during inflation and deflation, particularly after a change of body position.
Do not wrap the cuff on the following parts:
- An upper arm on which intravenous drip or a blood transfusion is being performed.
- An upper arm on which an SpO sensor, IBP catheter, or other instrument is attached.
2
- An upper arm with a shunt for hemodialysis
If measuring blood pressure with the cuff wrapped on the arm on the side of the body where a
mastectomy was performed, check the patient’s condition.
Caution
NIBP measurement should be performed on the upper arm.
During NIBP, stop excessive body movement by the patient and minimize trembling.
If a doctor has indicated that the patient has hemorrhagic diathesis or hypercoagulability, check the
condition of the arm after measurement.
Use the appropriate cuff size to ensure correct measurements. If too large a cuff is used, the
measured blood pressure value tends to be lower than the actual blood pressure. If too small a cuff is
used, the measured blood pressure value tends to be higher.
Before and during measurement, verify that none of the following apply to the patient:
- The part where the cuff is wrapped is at a different height than the heart.
(A difference of 10 cm (4 inches) in height may cause a variation in the blood pressure value of up
to 7 or 8 mmHg.)
- Body movement or conversing during measurement.
- Cuff wrapped over thick clothing.
- Pressure on the arm due to a rolled up sleeve.
In the case of a cuff for adults, the cuff should be wrapped to a tightness that allows two fingers to be
inserted in between the cuff and the arm.
The accuracy of a flashing measurement value that is out of the measurement range cannot be
guaranteed. Always check the patient’s condition before deciding what steps to take.
Do not use the cuff if it is damaged or has holes.
Only OMRON GS CUFF can be used with this device.
8
Download from Www.Somanuals.com. All Manuals Search And Download.
Note:
Setup
• Read and understand the manual for each optional accessory. This manual does not contain
cautionary information for optional accessory.
• Exercise caution with the cables and arrange so that the patient does not become entangled or
bound.
Before use / during use
• Check the following after turning on the power:
- No smoke, abnormal odor, or abnormal noise is emitted.
- Press each button and verify that it operates.
- For functions that cause icons to light or flash, verify that the icons light or flash (page 13).
- Measurement can be performed normally, and measurement error is within the tolerance value.
• If the screen is not displayed normally, do not use the unit.
• When recycling or disposing of parts (including batteries) of the unit, follow local government rules
and regulations.
Cleaning
• For cleaning, see page 24.
Rechargeable battery
• To prevent accidents, keep batteries out of reach of infants and small children.
• If you sense that something is wrong with a battery, immediately move it to a safe location and contact
the administrator responsible for the unit or OMRON HEALTHCARE.
• If the battery voltage is low, operation by battery may not be possible.
Non-Invasive Blood Pressure (NIBP) measurement
• If the patient has acute inflammation, a pyogenic ailment, or an external wound at the location where
the cuff is to be wrapped, follow the instructions of the doctor.
• Non-Invasive Blood Pressure measurement (NIBP) is performed by compressing the upper arm.
Some people may experience intense pain, or transient spotting caused by subcutaneous
hemorrhaging (bruising) may appear. The spotting will disappear with time; however, it may be
appropriate to inform patients for whom this may be a concern that spotting sometimes occurs, and if
necessary, refrain from measurement.
• To measure correctly, it is recommended that the patient relax and not talk during measurement.
• To measure correctly, it is recommended that the patient rest quietly for 5 minutes before
measurement.
EN
9
Download from Www.Somanuals.com. All Manuals Search And Download.
Using the Unit
Components of the Product
Before using the unit, make sure that no accessories are missing and that the unit and accessories are
not damaged. If an accessory is missing or there is damage, please contact OMRON HEALTHCARE.
Main unit
Standard Medical Accessories
• GS CUFF L (32 - 42 cm)/M (22 - 32 cm)
• AC adapter
Others
• Battery pack
• Instruction Manual (this paper)
Options
Optional Accessory
Battery pack
HXA-BAT-2000
(9065797-0)
10
Download from Www.Somanuals.com. All Manuals Search And Download.
Optional Medical Accessories
(within the scope of EC Medical Device Directive 93/42/EEC)
GS CUFF XL
GS CUFF L
GS CUFF M
GS CUFF S
HXA-GCUFF-XLLB
(9065802-0)
HXA-GCUFF-LLB
(9065798-9)
HXA-GCUFF-MLB
(9065799-7)
HXA-GCUFF-SLB
(9065800-4)
CUFF HOSE No.1
(3.5m)
HBP-CUFFH-BLU35
(9968171-8)
CUFF HOSE No.2
(1.5m)
HBP-CUFFF-BLU15
(9968172-6)
AC adapter*
AC ADAPTER-E1600
(9063658-2)
GS CUFF SS
HXA-GCUFF-SSLB
(9065801-2)
* UK plug type
AC ADAPTER-UK1600
(9994843-9)
Caution
Only OMRON GS CUFF can be used with this device.
EN
Features of the Product
The OMRON HBP-1300 is an affordable professional blood pressure unit that is clinically proven
accurate and provides fast, reliable results and is easy to use.
Key features, benefits, look
• 5 cuffs available - (SS: 12 to 18 cm, S: 17 to 22 cm, M: 22 to 32 cm, L: 32 to 42 cm, XL: 42 to 50 cm)
(12 to 50 cm arm circumference range)
• Designed to be used on a table
• Motion stop function (When body movement is detected, the device stops deflation for 5 seconds.)
• Irregular pulse Indicator - Helps identify changes in heart rate, rhythm, or pulse which may be caused
by heart disease or other serious health issues.
• Inflation Pressure Setting - 4 options: Auto, 220 mmHg, 250 mmHg, 280 mmHg
• Last reading display function
• Auto-Off
• Custom Rechargeable battery
• Large, easy-to-read backlight display
11
Download from Www.Somanuals.com. All Manuals Search And Download.
Features / Functions of Unit
Front and back of unit
4
6
5
2
3
1
7
Power ON / Measure blood pressure
Press when the power is off to turn on the power and start
blood pressure measurement. During inflation or
measurement, press to stop.
[START/STOP] button
(Power ON/OFF)
1
Power OFF
Hold down at least 3 seconds to turn off the power.
Press to enter “Menu Mode” and configure various settings.
If held down at least 3 seconds when a measurement result
is displayed, the measured data is cleared without being
stored in memory.
2
3
[MODE] button
Displays the system settings (page 16) and the last reading
(page 23). When using “Auscultation Mode”, inflates and
deflates the cuff.
[
] [
] button
4
5
6
7
Alarm lamp
NIBP connector
DC jack
Lights up or flashes when an alarm occurs (page 28).
Connects the cuff tube.
Connects the AC adapter.
Battery cover
Open to install or replace the battery.
Meaning of the Symbols
Symbol Description
Symbol Description
This shows the Type BF applied part.
Consult the instruction manual.
Class II (AC Adapter)
12
Download from Www.Somanuals.com. All Manuals Search And Download.
LCD Display
1
2
4
3
8
5
6
7
9
10
1
2
3
SYS
Displays systolic blood pressure.
Displays diastolic blood pressure.
Displays the pulse rate.
DIA
Pulse
Pulse
synchronization icon
Flashes in synchronization with the pulse during
measurement.
4
5
6
Lights in the measurement result display and memory
display if the pulse wave interval was irregular or there
was body movement during measurement (page 23).
Irregular pulse wave
icon
EN
Lights while the previous data are being displayed
(page 23).
Memory icon
Charge icon*
Flashes during charging. Solid colored light displays
when charging is finished.
7
8
9
Battery charge level
icon*
Displays how much charge the battery has (page 14).
Lights when “Auscultation Mode” is ON (page 16).
Auscultation icon
10
Inflation setting icon
The set initial Inflation pressure value lights (page 16).
* Only when the battery is installed.
13
Download from Www.Somanuals.com. All Manuals Search And Download.
Installing the Battery Pack
Warning
• If battery fluid comes in contact with the eye, immediately flush with copious amounts of water.
Do not rub. Seek medical attention immediately.
• Do not use the battery pack in any other device besides this unit. Do not throw into flame,
disassemble, or heat.
Caution
Do not short the positive and negative terminals of the battery pack with a wire or other metal object. If
battery fluid comes into contact with the skin or clothes, immediately rinse with water.
1. Make sure the AC adapter has been disconnected.
2. Remove the two screws from the top of the rear
unit cover, and remove the battery cover.
3. Connect the battery pack to the connector on the
monitor, and insert the battery pack into the
compartment so that the triangle mark on the
battery pack is aligned with the triangle mark on
the monitor.
4. Replace the battery cover and secure with screws.
Take care not to pinch the wires when replacing the cover.
5. Connect the AC adapter to the unit and charge the battery pack.
When using the battery pack for the first time, be sure to fully charge (about 4 hours) before using.
Battery life
• About 300 measurements are possible with one charge.
• A general guideline for replacement of the battery pack is about one year, however, the usage time
per charge may grow shorter depending on usage conditions. If you find that the usage time after
each charge has grown shorter and the
icon frequently appears, replace the battery pack.
Charging time
• Charging automatically starts after the AC adapter is connected.
When a new battery pack or a battery pack that has not been used for a long time is used, some time
may elapse before charging starts.
• The
icon flashes during charging.
• Charging is completed in about 4 hours.
Low battery
When the
icon starts flashing, promptly charge the battery pack.
When the icon changes to
battery.
, the battery is too low for blood pressure measurement. Charge the
Auto Power Off
• When the unit is powered by the battery pack, the power automatically turns off after the set time
elapses if you forget to turn off the power.
• When the unit is used with the AC adapter connected, the “Auto Power Off” setting does not operate.
14
Download from Www.Somanuals.com. All Manuals Search And Download.
Battery charging
State
LCD contents and operations
The icon flashes.
Icon
Charging
Fully charged
(Charging completed)
The icon is lit.
Problem with the
battery
Error message is displayed.
-
Battery level
Battery level
LCD contents and operations
Icon
The icon is lit.
The unit can be used.
Fully charged
The icon flashes (E40 error not displayed).
The unit can be used.
Battery level is 20%.
The icon flashes (E40 error displayed).
The unit cannot be used.
If the unit is being continuously used, the power will be
automatically turned off in 30 seconds.
-
Battery level is 5%.
Connecting the AC Adapter
AC power
Verify that the power outlet supplies the specified voltage and frequency (100 - 240 V AC, 50/60 Hz).
EN
Connect the AC adapter to the DC jack on the unit and the power outlet.
Note:
When the battery pack is installed
If there is no problem with the following.
- AC adapter
- DC jack
- Power outlet
- Battery
And the
icon does not flash, contact OMRON HEALTHCARE.
15
Download from Www.Somanuals.com. All Manuals Search And Download.
System Settings
The system settings are divided into two modes, “Menu Mode” and “Utility Mode”.
Menu Mode
“Menu Mode” allows you to configure the “Initial Inflation Pressure Value” and “Auscultation Mode” settings.
Initial Inflation
Pressure Value
Auscultation Mode
1. Press the [MODE] button.
The “Initial Inflation Pressure Value” setting screen appears.
2. Press the [ ] or [ ] button to change the setting.
• Press the [ ] button to change the setting in the order “280”, “250”, “220”, “AUTO”, “280”...
• Press the [ ] button to change the setting in the order “220”, “250”, “280”, “AUTO”, “220”...
When the “Initial Inflation Pressure Value” is set, inflation takes place at a fixed speed to the set
value and thus is quicker.
“AUTO” estimates the systolic blood pressure during inflation and automatically inflates the cuff to a
suitable value.
When using “220”, “250”, or “280” mmHg, select the value that is 30 to 40 mmHg higher than the
estimated systolic blood pressure.
3. When you have completed the “Initial Inflation Pressure Value” setting,
press the [MODE] button.
The “Auscultation Mode” settings screen appears.
“ON” or “OFF”
appears.
4. Press the [ ] or [ ] button to set to “ON” or “OFF”.
When set to “ON”, SYS and DIA can be recorded using auscultation measurement.
For information on auscultation measurement, see page 22.
5. When you have completed the “Auscultation Mode” setting, press the
[MODE] button.
“0” appears.
To start blood pressure measurement, press the [START/STOP] button.
Note:
• The body movement detection function is disabled while “Auscultation Mode” is in use.
• If the power is turned off, the settings revert to the factory settings.
“Initial Inflation Pressure Value” changes to “AUTO”, and “Auscultation Mode” changes to “OFF”.
16
Download from Www.Somanuals.com. All Manuals Search And Download.
Utility Mode
“Utility Mode” allows you to configure the “Auto Power Off” and “Pressure Accuracy Confirmation” setting.
1. Confirm that the device is switched off.
If the power is on, hold down the [START/STOP] button for at least 3 seconds to
turn off the power.
2. Hold down the [MODE] button until the “Auto Power Off”
setting screen appears.
The “Initial Inflation Pressure Value” setting screen appears, and changes to the
“Auto Power Off” setting screen appears.
3. Press the [ ] or [ ] button to change the auto power off
setting.
“5” or “10”
appears.
The “5 min.” or “10 min.” setting is entered.
When the unit is powered by the battery pack and is not used for the set time (“5 min.” or “10 min.”),
the power automatically turns off to save battery power.
If a mid-priority alarm other than a low battery error (E40 error) has occurred, the power does not
turn off automatically.
4.When you have completed the “Auto Power Off” setting, press the [MODE]
button.
EN
The “Pressure Accuracy Confirmation” screen appears. “0” appears.
“0” appears.
5. Check the Pressure Accuracy.
Add pressure externally as explained on page 24.
Compare with the displayed value and verify that there is no problem.
6. When you have completed “Pressure Accuracy Confirmation”, press the
[START/STOP] button.
The power turns off.
17
Download from Www.Somanuals.com. All Manuals Search And Download.
Non-Invasive Blood Pressure (NIBP) Measurement
Non-Invasive Pressure Measurement Principles
Oscillometric method
The beat in the pulsation generated by the contraction of the heart is captured as the pressure inside
the cuff to measure the blood pressure. If the cuff wrapped around the upper arm is pressurized
sufficiently, the blood flow stops, but the beat of the pulsation is present and the pressure inside the cuff
receives this and oscillates. Next, as the pressure inside the cuff gradually decreases, the oscillation of
the pressure within the cuff gradually increases and reaches a peak. As the pressure within the cuff
decreases further, the oscillation decreases from its peak.
The pressure within the cuff and the relationship with the increase and decrease of the oscillation within
the cuff in this series of processes are stored into memory, calculations are carried out, and the blood
pressure value is determined.
The pressure within the cuff when the oscillation increases drastically is the systolic pressure and the
pressure within the cuff when the oscillation decreases drastically is the diastolic pressure. Also, the
pressure within the cuff when the oscillation peaks is taken as the average pulsation pressure.
The oscillometric method does not determine the blood pressure value instantly like a microphone type
automatic blood pressure gauge with the auscultation method, but rather determines it from the series
of change curves as explained above. Therefore, it is not easily affected by external noise, an electric
scalpel or other electro surgical instruments.
KOROTKOV SOUNDS
CUFF
PRESSURE
OSCILLATIONS IN CUFF PRESSURE
RADIAL PULSE
5 SEC
Comparison between the auscultatory, oscillometric and
palpatory methods of measuring blood pressure.
L.A. Geddes,
“The Direct and Indirect Measurement of Blood Pressure”, Year Book Medical Publishers, Inc. 1970
18
Download from Www.Somanuals.com. All Manuals Search And Download.
Cuff Selection and Connection
Selecting the cuff
Warning
If a cuff is used on a patient with an infection, treat the cuff as medical waste, or disinfect before
reuse.
Caution
• Do not use the cuff if it is damaged or has holes.
• Use the appropriate cuff size to ensure correct measurements. If a cuff that is too large is used, the
measured blood pressure value tends to be lower than the actual blood pressure. If a cuff that is too
small is used, the measured blood pressure value tends to be higher.
Note:
It is important to use the correct sized cuff for a patient in order to get an accurate reading.
Measure the circumference of the patient’s arm and select the cuff size
that is appropriate for the circumference.
Select the cuff that is suitable for the patient from the cuffs below.
Arm circumference
Cuff name
(cm)
(inch)
17-20
13-17
9-13
7-9
GS CUFF XL*
GS CUFF L
42 - 50
32 - 42
22 - 32
17 - 22
12 - 18
GS CUFF M
GS CUFF S*
GS CUFF SS*
5-7
* Available as an optional accessory.
EN
Connecting the cuff
Connect the cuff tube to the NIBP connector on the unit and turn
clockwise to lock.
Caution
Only OMRON GS CUFF can be used with this device.
Note:
• If the cuff tube is too short, the optional 3.5 m or 1.5 m CUFF HOSE can be connected to lengthen the
cuff tube. Do not extend other than the optional 3.5 m or 1.5 m CUFF HOSE, as this will affect
measurement accuracy.
• Make sure the connections are tight.
19
Download from Www.Somanuals.com. All Manuals Search And Download.
Applying the Cuff to the Patient
1. Wrap on a bare arm or over thin clothing.
Wrap the cuff on a bare arm or over thin clothing. Thick clothing or a rolled up sleeve will cause
inaccurate blood pressure measurements.
The device can be used on either the right or left arm.
2. Align the artery mark “INDEX ARTERY” with the brachial artery.
Run the cuff tube out the peripheral side with no bends (the brachial artery is on the inner side of the
patient’s upper arm).
Make sure that “INDEX ARTERY” is within the “RANGE” and the lower edge of the cuff is 1 to 2 cm
from the inner side of the elbow joint.
If “INDEX ARTERY” is outside the “RANGE”, error in the blood pressure value will increase. In this
case, use a different cuff size.
As a guideline for the tightness of the cuff, It should be possible to insert about two fingers under the
cuff.
1.
2.
3.
1 to 2 cm
3. During measurement, keep the brachial artery on which the cuff is
wrapped at the same height as the right atrium of the heart.
20
Download from Www.Somanuals.com. All Manuals Search And Download.
Caution
Make sure the cuff is wrapped in the correct arm position and is at the same height as the heart.
A difference of 10 cm (4 inches) in height may cause a variation in the blood pressure value of up to
7 - 8 mmHg.
Note:
• If measurement is difficult due to arrhythmia, use a different blood pressure measurement method.
• If the patient has acute inflammation, a pyogenic ailment, or an external wound at the location where
the cuff is to be wrapped, follow the instructions of the doctor.
• Non-Invasive Blood Pressure (NIBP) measurement is performed by compressing the upper arm.
Some people may experience intense pain, or transient spotting caused by subcutaneous hemorrhaging
may appear. The spotting will disappear over time, however, if it is possible that this will disturb the patient,
try the following technique:
- Wrap a thin towel or cloth (one layer) under the cuff.
If the towel or cloth is too thick, there will be insufficient cuff compression and the blood pressure value
will measure high.
• If the patient moves or the cuff is touched, this may be falsely detected as a pulse and over-inflation
will occur.
• Do not inflate the cuff when it is not wrapped on the upper arm. This may damage the cuff.
Taking the Measurement in “Manual Mode”
1. Press the [START/STOP] button.
Blood pressure measurement is performed once.
2. The measurement results are displayed.
If a measurement value is outside the corresponding range below, the value will flash.
SYS: 59 mmHg or less, or 251 mmHg or higher.
DIA: 39 mmHg or less, or 201 mmHg or higher.
PULSE: 39 bpm or less, or 201 bpm or higher.
ꢀ Normal measurement
ꢀ Measurement error /
failure
EN
Caution
The accuracy of a flashing measurement value that is outside the measurement range is not
guaranteed. Always check the patient’s condition before deciding what steps to take.
Note:
If inflation is insufficient, inflation may restart automatically while measurement is in progress.
21
Download from Www.Somanuals.com. All Manuals Search And Download.
Irregular pulse wave detection function
If the pulse wave interval becomes irregular during measurement, the irregular pulse wave detection
icon will light to notify you.
Body movement detection function
If body movement is detected during measurement, deflation stops for 5 seconds. The irregular pulse
wave icon appears in the measurement result display.
ꢀ Deflation stopped
After 5 seconds, measurement resumes, and an attempt is made to complete measurement in one
cycle.
Taking the Measurement in “Auscultation Mode”
Use a stethoscope to perform measurement.
When set to “ON”, SYS and DIA can be recorded using auscultation measurement.
To determine SYS and DIA during measurement, press the [MODE] button.
SYS is registered the first time and DIA is registered the second time you press the [MODE] button
during deflation.
After DIA is determined, the cuff rapidly deflates and SYS and DIA are displayed as the measurement
results.
The pulse rate does not appear in the measurement result display.
During deflation, the [ ] button can be held down to re-inflate, or the [ ] button can be held down to
deflate faster.
Stored data that was measured in “Auscultation Mode” is displayed when the auscultation icon is lit.
ꢀ During re-inflation
Note:
The body movement detection function is disabled while “Auscultation Mode” is in use.
For Auscultation Mode settings, see page 16.
22
Download from Www.Somanuals.com. All Manuals Search And Download.
Stopping the Measurement
To stop measurement while measurement is in progress, press [START/STOP] button.
Displaying Last Reading
The previous measurement value (systolic blood pressure, diastolic blood pressure, and pulse rate)
and whether or not an irregular pulse wave was detected can be displayed. In the case of auscultation
measurement, the auscultation icon appears.
Press the [ ] or [ ] button to display the previous data. This function is also available when the
device is switched off.
If an irregular pulse wave or
body movement was detected,
the irregular pulse wave icon
Lights while the previous
data are being displayed.
appears.
When the data was measured
in “Auscultation Mode”, the
auscultation icon appears.
If held down at least 3 seconds when a measurement result is
displayed, the measured data is cleared without being stored.
Note:
When the monitor is left idle for one minute, the backlight will disappear.
EN
23
Download from Www.Somanuals.com. All Manuals Search And Download.
Maintenance
Maintenance Inspection and Safety Management
The HBP-1300 must be maintained to ensure functionality and to secure the safety of patients and operators.
Daily checks and maintenance should be performed by the operator. (page 25)
In addition, qualified personnel are necessary to maintain the performance and the safety, and to conduct
periodic inspections. We recommend that the verification test be performed at least once a year.
Example of connection for pressure accuracy confirmation:
1. Display the pressure accuracy confirmation screen as explained in
“Utility Mode” on page 17.
Display “0” in the pressure accuracy confirmation screen.
Display the pressure value.
2. Connect the blood pressure monitor, the calibrated reference pressure
gauge, and the cuff and inflation bulb.
3. Check the pressure value of the blood pressure monitor and the
pressure value of the calibrated reference pressure gauge.
Note:
• Make sure that the blood pressure monitor reading is within 3 mmHg compared to the calibrated
reference pressure gauge.
• To rapidly deflate the cuff, press the [ ] button.
To repeat “Pressure Accuracy Confirmation”, turn off the power and repeat the procedure from
step 1 in “Utility Mode” on page 17.
Cleaning of the Device
Cleaning and disinfecting should be performed in accordance with your facility’s infection control
practice.
Surface cleaning
Wipe with a cloth that has been moistened with isopropyl alcohol diluted to 50 v/v%, or ethyl alcohol
(disinfection alcohol) diluted to 80 v/v% or less and wrung out.
Do not wipe the DC jack or allow it to become wet.
Removing dust
Use a moistened cotton bud to remove dust that has accumulated on the vent ports.
Service
The device requires no routine service other than cleaning, and visually checking the cuffs, tubing, etc.
Caution
• Do not sterilize by autoclave or gas sterilization (EOG, formaldehyde gas, high-concentration ozone, etc.).
• If using an antiseptic solution for cleaning, follow the instructions of the manufacturer.
Accessory Care
Non-Invasive Blood Pressure Measurement (NIBP)
Cuff
Wipe clean on the surface of the cuff with a cloth moistened with a 70 v/v% dilution of isopropyl alcohol,
or a 80 v/v% or less dilution of disinfection ethanol (ethyl alcohol).
Do not allow any liquids inside the cuff. If a liquid gets in the cuff, dry the inside well.
24
Download from Www.Somanuals.com. All Manuals Search And Download.
Check before Use
Before conducting safety checks, be sure to implement the items in the “Cleaning of the Device” and
“Accessory Care” sections. (page 24)
Before turning on the power
Before turning on the power, check for the following
External appearance
ꢀ The device or accessories are not deformed due to falling or other impact.
ꢀ The device is not dirty.
ꢀ The device is not wet.
AC adapter
ꢀ The AC adapter is firmly connected to the connector on the device.
ꢀ There are no heavy objects lying on the AC adapter cable.
ꢀ The AC adapter cable is not damaged (core-wire exposure, breaks, etc.).
When turning on the power
When turning on the power, check the following
Display/lamp
ꢀ When the [START/STOP] button is pressed to turn on the power, the screen below appears and the
alarm lamp lights (page 13).
The alarm lamp is lit up.
EN
After turning on the power
After turning on the power, check for the following
External appearance
ꢀ There is no smoke or odor coming from the device.
ꢀ The device is not making any unusual noises.
Buttons
ꢀ Press each button and check that it works.
Non-invasive blood pressure (NIBP)
ꢀ Make sure that a suitable OMRON GS CUFF is attached (one that fits the circumference of the
patient’s arm).
ꢀ The cuff tube is firmly connected.
ꢀ The person checking the cuff should wrap the cuff around arm, perform cuff measurement and check
to see that blood pressure is in the vicinity of normal measurements.
ꢀ While measurement is in progress, bend the relevant arm and move body to halt discharge and
during this halt check that cuff pressure does not drop.
25
Download from Www.Somanuals.com. All Manuals Search And Download.
Troubleshooting
The power does not turn on
Cause
Solution
If the unit is being powered by the battery, the battery
is not installed or the charge is depleted.
Insert battery or replace with a new battery (page 14).
Disconnect the AC adapter, remove the battery, and
contact OMRON HEALTHCARE.
Internal part failure
• Check if the AC adapter is disconnected or the connection is loose.
• Check if the AC adapter or battery has failed.
The unit display does not operate
Cause / solution
Stop using the unit and contact OMRON HEALTHCARE.
The unit becomes hot
Cause
Solution
An object is on top of the unit or right next to the unit. Keep the area around the unit free of objects.
If the unit becomes too hot to be touched, there may be a problem in the unit. Turn off the unit power,
disconnect the AC adapter, remove the battery, and contact OMRON HEALTHCARE.
The unit is connected to a power outlet, but it runs on the the battery pack
Cause / solution
If AC power cannot be supplied, the unit will operate by battery only.
• Check if the AC adapter is properly connected to the unit.
• Check if AC adapter is connected to a power outlet.
• Check that the electrical outlet is working by connecting a different device to the same power outlet.
The cuff does not inflate when the [START/STOP] button is pressed
Cause
Solution
Loose cuff tube connection.
There is an air leak in the cuff.
If pressure is displayed, the cuff tube is bent.
Check the connection.
Replace the cuff.
Make sure no part of the cuff tube is bent.
Measurement was not possible
Cause / solution
Check the patient by palpation or other method.
After checking the patient, check the error code and see “List of Error Codes” (page 28) for Non Invasive
Blood Pressure (NIBP) measurement.
26
Download from Www.Somanuals.com. All Manuals Search And Download.
Abnormal measurement value
Cause / solution
The causes below are possible. Check the patient by palpation and then repeat measurement.
• Body movement (chills or other trembling)
• Arrhythmia.
• Noise in the cuff
- A nearby person touched the patient.
- Cardiac massage was being performed.
The measurement value is questionable
Cause
Solution
Deflates quickly
Check for a loose cuff connection.
Simultaneously perform measurement with a stethoscope.
Place the stethoscope and listen while viewing the pressure display of the
manometer.
Stethoscope
Blood pressure may vary widely due to physiological effects.
The causes below are possible.
• Emotional excitement or agitation
- Pain due to cuff wrapping
- White coat hypertension
• Cuff size or wrapping method not correct
• Cuff wrapping position on upper arm not at the same height as the heart
• Patient’s blood pressure not stable due to pulsus alternans, respiratory changes, or other reason
Measure circumference or patient’s arm and ensure
Incorrect cuff size used.
correct sized cuff is used.
Ensure cuff is applied to a bare arm, or very thin
EN
Cuff wrapped over thick clothing.
clothing.
Ensure patient is seated, feet flat on the floor, cuff at
Patient not seated properly.
heart level.
Ensure before measurement taken, patient has not
Patient ate, drank, or exerted themselves recently.
had food, caffeinated or alcohol beverages, or
exerted/exercised in the last 30 minutes.
27
Download from Www.Somanuals.com. All Manuals Search And Download.
List of Error Codes
The alarm lamp flashes when a medium-priority alarm occurs, and lights steadily when a low-priority
alarm occurs.
To clear an alarm, press any button.
ꢀ If a low-priority alarm and a medium-priority alarm occur at the same time, the medium-priority alarm
is displayed.
If the alarms are the same priority level, the error code of the alarm that occurred first is displayed.
However, to prevent battery consumption when powered by battery, a low battery error (E40 error) is
always given priority.
ꢀ Example: E2
SYSTEM
Error
code
Priority
Description
Points to check
E9
Medium Internal hardware error
Contact OMRON HEALTHCARE.
NIBP
Error
code
Priority
Description
Points to check
The cuff tube is not connected
Air is leaking from the cuff.
Firmly connect the cuff tube.
E1
Low
Replace with an OMRON GS CUFF that
does not leak.
Did not inflate properly because the arm Have the patient not move the arm or body,
or body moved during measurement.
and repeat measurement.
Moved body or arm during
measurement, or talked.
Have the patient not talk or move, and
repeat measurement.
The cuff is not applied correctly.
Correctly apply the cuff.
The sleeve is rolled up and is
compressing the arm.
Remove the garment and re-wrap the cuff.
E2
Low
The measurement time exceeds the
expected time, so the measurement was
ended in order to avoid patient discomfort.
There is a possibility that measurement is
being repeated over and over due to air
leaking from the cuff.
Measurement time has exceeded
specified time.
Specified time: 165 seconds
28
Download from Www.Somanuals.com. All Manuals Search And Download.
Other problems
Error
Priority
code
Cause
Solution
When inflating in “Auscultation Mode”,
release the button when the pressure
reaches the desired value.
Inflated the cuff to 300 mmHg or more
during inflation in “Auscultation Mode”.
If this occurs during measurement, repeat
measurement.
E3
Low
Over-inflation occurs
If this occurs when not performing
measurement, contact OMRON
HEALTHCARE.
Recharge the battery, or replace with a new
battery. (page 14)
E40
E41
Medium The battery is depleted.
Medium Battery failed to charge.
Try charging again. If continues to fail,
replace with a new battery. (page 14)
Replace the battery with a new battery. If the
error continues, contact OMRON
HEALTHCARE.
E42
Medium Battery voltage error
Disposal
Description
As there is a risk of environmental pollution, follow your applicable national and local legal regulations
regarding disposal or recycling of this equipment and batteries.
The main constituents of each part are listed in the table below. As there is a risk of infection, do not
recycle patient attachments such as cuffs, but dispose of them as instructed by your facility’s
procedures and applicable regulations.
Item
Parts
Material
Box
Cushion
Bag
Cardboard
Cardboard
PE
Package
EN
Enclosure
Internal parts
Outer tube
Cell batteries
Internal parts
ABS, PC , SR
General electronic components
PVC
Nickel-metal hydride
General electronic components
Main unit and
accessories
Battery pack
29
Download from Www.Somanuals.com. All Manuals Search And Download.
Specifications
Factory Default Settings
Factory default settings and backup are as shown below.
Backup
: Setting is retained even if the power is interrupted.
: Reverts to factory default setting if the power is turned off.
Setting Pressure Value
Initial Inflation Pressure Value
Auscultation Mode
Settings
AUTO, 220, 250, 280
ON, OFF
Factory Setting
AUTO
Backup
OFF
Auto Power Off
5 min, 10 min
5 min
Technical Specifications: HBP-1300
Main unit
Measurement Parameter
Dimension
NIBP, PR
Main unit: 123 x 201 x 99 (mm) 4.84 x 7.91 x 3.90 (inch) (W x H x D)
AC adapter: 46 x 66 x 37 (mm) 1.81 x 2.60 x 1.46 (inch) (W x H x D)
Battery: 54 x 43.5 x 15.4 (mm) 2.13 x 1.71 x 0.61 (inch) (W x H x D)
Weight
Main unit: Approx. 0.52 kg (not including accessories and options)
AC adapter: Approx. 0.2 kg
Battery: Approx. 0.1 kg
Display
7 segment LCD
Protection Class
Class II (AC Adapter)
Internal powered equipment (when operating with battery only)
Degree of Protection
Type BF
MDD Classification
Power supply
AC adapter
Class II a
Input voltage range: AC 100 V to 240 V
Frequency: 50/60 Hz
Output voltage range: DC 6 V 5%
Rated output current: 1.6 A
Rechargeable battery
Type: 3.6 V, 1900 mAh
Number of operation cycles when fully charged: 300
Measurement conditions
• New battery fully charged
• Ambient temperature of 23°C (73.4°F)
• Using M-size cuff
• SYS 120 / DIA 80 / PR 60 (Inflation setting: AUTO)
• One 5-minute cycle consisting of “cuff measurement time + wait time”
30
Download from Www.Somanuals.com. All Manuals Search And Download.
Environmental Conditions
Operational temperature and
humidity
Temperature range: 10 to 40°C (50 to 104°F)
Humidity range: 30 to 85%RH (not condensed)
Atmospheric pressure: 700 to 1060hPa
Storage and transportation
Temperature range: -20 to 60°C (-4 to 140°F)
Humidity range: 10 to 95%RH (not condensed)
Atmospheric pressure: 500 to 1060hPa
Non-Invasive Blood Pressure (NIBP)
Measurement technology
Measurement method
Oscillometric
Dynamic Linear Deflation method
0 to 300 mmHg
Pressure display range
Pressure display accuracy
NIBP measurement range
Within 3 mmHg
SYS 60 to 250 mmHg
DIA 40 to 200 mmHg
Pulse rate 40 to 200 /min
NIBP accuracy*
Maximum mean error within 5mmHg
Maximum standard deviation within 8mmHg
Pulse rate accuracy
Reference standard:
Within 5 % of reading
EN1060-1:1995+A2:2009
EN1060-3:1997+A2:2009
ISO81060-1:2007
* Comparison with auscultation method performed by a trained professional.
DIA determined by the auscultation method is “K5”.
Note: Specifications may be changed without prior notice.
EN
This blood pressure monitor fulfils the requirements of the EC directive 93/42/EEC (Medical Device
Directive). It also conforms to the European standard EN 1060, Non-invasive Sphygmomanometers
Part 1: General Requirements and Part 3: Additional Requirements for Electromechanical Blood
Pressure Measuring Systems.
31
Download from Www.Somanuals.com. All Manuals Search And Download.
Important information regarding Electro Magnetic Compatibility (EMC)
With the increased number of electronic devices such as PC’s and mobile (cellular) telephones,
medical devices in use may be susceptible to electromagnetic interference from other devices.
Electromagnetic interference may result in incorrect operation of the medical device and create a
potentially unsafe situation.
Medical devices should also not interfere with other devices.
In order to regulate the requirements for EMC (Electro Magnetic Compatibility) with the aim to
prevent unsafe product situations, the EN60601-1-2:2007 standard has been implemented. This
standard defines the levels of immunity to electromagnetic interferences as well as maximum levels
of electromagnetic emissions for medical devices.
This medical device manufactured by OMRON HEALTHCARE conforms to this EN60601-1-2:2007
standard for both immunity and emissions.
Nevertheless, special precautions need to be observed:
• Do not use mobile (cellular) telephones and other devices, which generate strong electrical or
electromagnetic fields, near the medical device. This may result in incorrect operation of the unit
and create a potentially unsafe situation. Recommendation is to keep a minimum distance of 7 m.
Verify correct operation of the device in case the distance is shorter.
Further documentation in accordance with EN60601-1-2:2007 is available within this manual, refer to
section “Manufacturer’s Declaration”.
Correct Disposal of This Product
(Waste Electrical & Electronic Equipment)
This marking shown on the product or its literature, indicates that it should not be
disposed with other household wastes at the end of its working life. To prevent possible
harm to the environment or human health from uncontrolled waste disposal, please
separate this product from other types of wastes and recycle it responsibly to promote
the sustainable reuse of material resources.
Household users should contact either the retailer where they purchased this product, or their local
government office, for details of where and how they can take this item for environmentally safe
recycling.
Business users should contact their supplier and check the terms and conditions of the purchase
contract. This product should not be mixed with other commercial wastes for disposal.
This product does not contain any hazardous substances.
Disposal of used batteries should be carried out in accordance with the national regulations for the
disposal of batteries.
32
Download from Www.Somanuals.com. All Manuals Search And Download.
Manufacturer’s Declaration
The HBP-1300 is intended for use in the electromagnetic environment specified below.
The customer or the user of the HBP-1300 should assure that it is used in such an environment.
Electromagnetic Emissions:
(IEC60601-1-2)
Emission Test
RF emission
CISPR 11
Compliance
Electromagnetic Environment
Group 1
The HBP-1300 uses RF energy only for internal functions.
Therefore, this RF emission is extremely weak and there is
little chance of it creating any kind of interference
whatsoever with nearby electronic equipment.
RF emissions
CISPR 11
Class B
Class A
Complies
The HBP-1300 is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low voltage power supply network
that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker
IEC 61000-3-3
Electromagnetic Immunity:
(IEC60601-1-2)
IEC60601-1-2
test level
Compliance
level
Electromagnetic
environment - guidance
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
6 kV contact
8 kV air
6 kV contact
8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30 %.
Electric fast transient/
burst
2 kV for
2 kV for
Mains power quality should be that of
power supply lines power supply lines a typical commercial or hospital
IEC 61000-4-4
1 kV for input/
1 kV for input/
environment.
output lines
output lines
EN
Surge
IEC 61000-4-5
1 kV
differential mode
2 kV
1 kV
differential mode
2 kV
Mains power quality should be that of
a typical commercial or hospital
environment.
common mode
common mode
Voltage dips, short
interruptions and
voltage variations on
power supply input lines
IEC 61000-4-11
<5 % U
for 0.5 cycle
<5 % U
for 0.5 cycle
Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the
HBP-1300 requires continued
operation during power mains
interruptions, it is recommended that
the HBP-1300 be powered from an
uninterruptible power supply or a
battery.
T
T
40 % U
40 % U
T
T
for 5 cycles
for 5 cycles
70 % U
70 % U
T
T
for 25 cycles
for 25 cycles
<5 % U
for 5 sec.
<5 % U
for 5 sec.
T
T
Power frequency
(50/60 Hz) magnetic
field
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.
IEC 61000-4-8
Note: U is the a.c. mains voltage prior to application of the test level.
T
33
Download from Www.Somanuals.com. All Manuals Search And Download.
IEC60601-1-2
test level
Compliance
level
Electromagnetic
Immunity test
environment - guidance
Portable and mobile RF communications
equipment should be used no closer to any part
of the HBP-1300, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency of
the transmitter.
Recommend separation distance
P
P
P
d = 1.2
d = 1.2
d = 2.3
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
3 Vrms
Conducted RF
IEC 61000-4-6
150 kHz to
80 MHz
3 Vrms
3 V/m
80% AM (2Hz)
where P is the maximum output power rating
of the transmitter in watts (W) according to he
transmitter manufacturer and d is the recom-
mended separation distance in meters (m).
3 V/m
Radiated RF
80 MHz to
2.5 GHz
Field strengths from fixed RF transmitters as
determined by an electromagnetic site
IEC 61000-4-3
80% AM (2Hz)
a
survey , should be less than the compliance
b
level in each frequency range .
Interference may occur in the vicinity of
equipment marked with the following symbol:
Note1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due
to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the HBP-1300 is used exceeds the applicable RF
compliance level above, the HBP-1300 should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as reorienting
or relocating the HBP-1300.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
34
Download from Www.Somanuals.com. All Manuals Search And Download.
Recommended Separation Distances:
Recommended separation distance between portable and mobile RF communications equipment
and the HBP-1300
The HBP-1300 is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the HBP-1300 can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters)
and the HBP-1300 as recommended below, according to the maximum output power of the communications
equipment.
Separation distance according to frequency of transmitter
m
Rated maximum output
power of transmitter
(W)
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
P
P
P
d = 1.2
0.12
0.38
1.2
d = 1.2
0.12
0.38
1.2
d = 2.3
0.23
0.73
2.3
0.01
0.1
1
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in metres (m) can be determined using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
Note1: At 80MHz and 800MHz, the separation distance for the higher frequency range applies
Note2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
EN
35
Download from Www.Somanuals.com. All Manuals Search And Download.
Manufacturer
OMRON HEALTHCARE Co., Ltd.
53, Kunotsubo, Terado-cho, Muko, Kyoto, 617-0002 JAPAN
EU-representative OMRON HEALTHCARE EUROPE B.V.
Scorpius 33, 2132 LR Hoofddorp, THE NETHERLANDS
OMRON (DALIAN) CO., LTD.
Production facility
Dalian, CHINA
OMRON HEALTHCARE UK LTD.
Opal Drive, Fox Milne, Milton Keynes, MK15 0DG, U.K.
OMRON MEDIZINTECHNIK HANDELSGESELLSCHAFT mbH
Gottlieb-Daimler-Strasse 10, 68165 Mannheim, GERMANY
Subsidiary
OMRON SANTÉ FRANCE SAS
14, rue de Lisbonne, 93561 Rosny-sous-Bois Cedex, FRANCE
Made in China
Download from Www.Somanuals.com. All Manuals Search And Download.
|