Lumiscope Blood Pressure Monitor 1131 User Manual

A U T O M A T I C U P P E R A R M  
Blood Pressure Monitor  
Model 1131  
Features:  
• Dual Person Memory Zones  
• 50 Memories for each zone  
• Average for Last 3 Memories  
• Date and Time Stamp  
• One Touch Operation  
Table of Contents:  
Introduction .................................. 2  
Name/Function of Each Parts ........ 2  
Quick Read Measuring Technology .. 3  
Preliminary Remarks ...................... 3  
About Blood Pressure..................... 3  
Blood Pressure Standard................ 4  
Blood Pressure Fluctuation............. 4  
Display Explanation ........................ 5  
Installing Battery............................ 6  
Applying the Cuff............................ 7  
Measurement Procedures............ 8,9  
Recalling Values from Memory...... 10  
Clearing Values from Memory ....... 11  
Time Adjustment .......................... 12  
Helpful Tips.................................. 13  
Troubleshooting............................ 14  
Cautionary Notes.......................... 15  
Specifications .............................. 16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Quick Read Measuring Technology  
This monitor uses trend-setting Quick Read Technology to detect your blood pressure. With one  
touch of a button, the device will automatically start to inflate and determine your blood pressure  
by analysis of your pressure oscillations. When the measurement is completed, the device  
will exhaust cuff pressure and display your systolic, diastolic, and pulse. Measurement time is  
significantly shortened, and user comfort is greatly enhanced.  
About Blood Pressure  
What is Blood Pressure?  
Blood pressure is the pressure exerted on the artery tube while blood flows through the arteries.  
The pressure measured when the heart contracts and sends blood out of the heart is systolic  
(highest). The pressure measured when the heart dilates with blood flowing back into the heart  
is called diastolic (lowest) blood pressure.  
Why Measure Your Blood Pressure?  
Among the various health problems today, problems associated with high blood pressure are by  
far the most common. High blood pressure's dangerously strong correlation with cardiovascular  
diseases has made measuring blood pressure a necessity for identifying those at risk.  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
Blood Pressure Standard  
The World Health Organization (WHO) has developed a blood pressure standard, according to  
which areas of low- and high-risk blood pressure are identified. This standard, however, is a  
general guideline and blood pressures vary between different people, age groups, etc.  
It is important that you consult with your physician regularly. Your physician will tell you your  
normal blood pressure range as well as the point at which you will be considered at risk.  
Blood Pressure Fluctuation  
Blood pressure fluctuates all the time!  
You should not be overtly worried if you encountered two or three measurements at high levels.  
Blood pressure changes over the month and even throughout the day. It is also influenced by  
season and temperature.  
4
Download from Www.Somanuals.com. All Manuals Search And Download.  
Display Explanation  
Display:  
Date/Time indicator  
Systolic Pressure  
Memory Zones  
Diastolic Pressure  
Pulse Rate  
Memory Number  
Icons:  
Memory Zone 1: Appears when the monitor is operated in memory zone 1  
Memory Zone 2: Appears when the monitor is operated in memory zone 2  
Weak Battery Mark: Appears when batteries should be replaced  
Pulse Mark: Shows the pulse rate per minute  
Memory Mark: Shows the number of stored measurements  
Error Mark: Occurs when a mistake was made during measurement  
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
Installing Battery  
1. Press down and lift the battery cover in the direction of the arrow to open the battery  
compartment.  
2. Install or replace 4 “AA” sized batteries in the battery compartment according to the  
indications inside the compartment.  
3. Replace the battery cover by clicking in the bottom hooks first, then push in the top end of the  
battery door.  
You need to replace the batteries when:  
1. low battery icon appears on display.  
2. the ON/OFF/START button is pushed and nothing appears on display.  
Caution:  
Batteries are hazardous waste.  
Do not dispose together with the household garbage.  
Using the AC adapter  
1.Connect the AC adapter with the AC adapter jack in the  
back of the unit.  
2.Plug the AC adapter into a standard wall socket.  
Caution:  
1. Please unload the batteries when operating  
with the AC mode for a longer period of time
Leaving the batteries in the compartment for  
a long time may cause leakage, which may  
lead to damage of the unit.  
2. No batteries are needed when operating with the AC adapter.  
6
Download from Www.Somanuals.com. All Manuals Search And Download.  
Applying the arm cuff  
1. Plug in the cuff connecting tube into the unit.  
2. Unwrap the arm cuff, leaving the “Index” end of the cuff through the D-ring of the cuff.  
3. Put your left arm through the cuff loop. The "OK" range indication should be positioned closer  
to yourself with the rubber tube pointing in the direction of your arm. Position the artery mark  
(
) over the main arteries (on the inside of your arm) in the upper arm.  
y
4. Turn your left palm upward  
and place the edge of the arm  
cuff  
at  
approximately  
2
to  
3
cm above the inner side of the elbow joint.  
Tighten the cuff by pulling the "Index" end  
of the cuff.  
5. If the Index line falls within the OK range,  
indicated at the edge of the cuff, this cuff  
suitable for your use. If the Index line falls out  
the OK range, you may need a special cuff  
contact our service department at  
800-672-8293 for more information regarding  
is  
of  
for  
or  
1-  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
Measurement Procedure  
1. Press the User-Switching key to select Memory Zone 1 or Memory Zone 2.  
or  
After a memory zone is selected, press th
measurement in the chosen memory zone.  
button to reset the monitor so it can start  
2. Press the  
button. All digits will light up, checking the display functions. The checking  
procedure will be completed after about 2 seconds.  
3. After all symbols appear,  
the  
display will show a blinking “0”. The monitor is now ready to measure and will automatically  
inflate the cuff slowly to start measurement.  
8
Download from Www.Somanuals.com. All Manuals Search And Download.  
Measurement Procedure  
4. When the measurement is completed, the cuff will automatically deflate. Systolic, diastolic  
and pulse will be shown simultaneously on the LCD screen. The measurement is then  
automatically stored into the pre-designated memory zone.  
Note:  
1. This monitor automatically switches off approximately 1 minute after last measurement.  
2. To interrupt the measurement, simply press any key. The cuff will deflate immediately after  
any key is pressed.  
3. Do not talk or move your arm or hand muscles during measurements.  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
Recalling Values from Memory  
1. The Monitor has two Memory Zones (1 and 2). Each zone  
can store up to 50 measurements.  
2. To read memory values from a selected Memory Zone,  
use the “User-Switching Key” to select a Memory Zone  
(1 or 2) from which you want to recall values from. Then,  
press the “M” key, and the values will be shown on the  
LCD display with an assigned memory sequence number.  
3. The monitor will display an average of the previous  
measurements first. Every new press of the “M” key  
calls for one set of prior values stored. Latest memory  
will be recalled first. The calculated average will be  
applied to a maximum of the last 3 memories.  
10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Clearing Values from Memory  
1. Press the User-Switching key to select the memory zone 1 or memory zone 2.  
2. Continue to press and hold the “M” key for approximately 5 seconds, and the data in the pre-  
designated memory zone will be erased automatically.  
Press “M” Key for  
5 seconds to clear all  
memories  
Note : The data in both memory zones (zone 1 and zone 2) will be erased if any of the batteries  
are removed, or if the AC adapter is disconnected from the unit or unplugged from the  
outlet. (It is suggested that users record their data on the Blood Pressure Log before  
removing or replacing batteries.)  
11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Time Adjustment  
1. To adjust the date/ time in the monitor, press the key . The display will show a blinking  
number showing the month.  
2. Change the month by pressing the “ ” Key. Each press will increase the number by one  
in a cycling manner. Press the Key. again to confirm the entry, and the screen will show a  
blinking number representing the date.  
3. Change the date, the hour, and the minute as described in Step 2 above, using the “ ” Key  
to change and the  
Set Key to confirm the entries.  
4. A “0” will reappear as the Blood Pressure Monitor is ready for measurement again.  
Press the  
Set-key  
blinking month  
increasing month  
confirm the entry  
12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Helpful Tips  
Here are a few helpful tips to help you obtain more accurate readings:  
• Do not measure your blood pressure immediately after consuming a large meal. Wait at least  
one hour after eating to take a measurement.  
• Do not smoke or drink alcohol before measuring your blood pressure.  
• You should not be physically tired or exhausted while taking a measurement.  
• It is important that you relax during measurement. Take 15-minute rest before a reading.  
• Do not take measurements if you are under stress or under tension.  
• Take your blood pressure at normal body temperature. If you are feeling cold or hot, wait a  
while before taking a measurement.  
• If the monitor is stored at very low temperature (near freezing), have it placed at a warm  
location for at least one hour before using it.  
• Wait about 5 minutes in between consecutive measurements.  
13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Troubleshooting  
If any abnormality should arise during use, please check the following points.  
Symptom  
Check Point  
Correction  
No display when the  
ON/OFF/START  
key is pressed  
Have the batteries run  
down?  
Replace them with four  
new batteries or use AC  
Adapter.  
Have the batteries’  
polarities been positioned  
incorrectly?  
Re-insert the batteries in  
the correct positions.  
EE mark shown  
on display or the  
blood pressure  
value is displayed  
excessively low  
(high)  
Is the cuff placed  
correctly?  
Wrap the cuff properly  
so that it is positioned  
correctly.  
Did you talk or move  
during measurement?  
Measure again quietly.  
Did you vigorously  
shake the cuff during  
measurement?  
Note: If the unit still does not work, return it to your dealer or contact our repair department.  
Under no circumstance should you disassemble and attempt to repair the unit by yourself.  
14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Cautionary Notes:  
1. The unit contains high-precision assemblies. Therefore, avoid extreme temperatures, humidity,  
and direct sunlight. Avoid dropping or strongly shocking the main unit, and protect it from  
dust.  
2. Clean the blood pressure monitor body and the cuff carefully with a slightly damp, soft cloth.  
Do not press. Do not wash the cuff or use chemical cleaner on it. Never use thinner, alcohol or  
petrol (gasoline) as cleaner.  
3. Leaky batteries can damage the unit. Remove the batteries when the unit is not used for a  
long time.  
4. The unit should not be operated by children so to avoid hazardous situations.  
5. If the unit is stored near freezing, allow it to acclimate at room temperature before use.  
6. Your Lumiscope Model 1131 is not field serviceable. You should not use any tool to open the  
device nor should you attempt to adjust anything inside the device. If you have any problems,  
please contact the store or the doctor from whom you purchased Lumiscope Model 1131 or  
contact the Lumiscope Repair Department at 1-800-672-8293.  
7. For users diagnosed with common arrhythmia (atrial or ventricular premature beats or atrial  
fibrillation), diabetes, poor circulation of blood, kidney problems, or for users who have  
suffered from stroke, the device might not be suitable for use.  
8. Do not use on unconscious patients.  
15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Specifications:  
Measurement Method  
Measurement Range  
Pressure Sensor  
Accuracy  
: Oscillometric  
: Pressure: 20~280mmHg; Pulse: 40~180 beats/ minute  
: Semi conductor  
: Pressure: ± 3mmHg; Pulse: ± 5% of reading  
: Pump Driven  
Inflation  
Deflation  
Memory capacity  
Auto-shut-off  
: Automatic Pressure Release Valve  
: 50 memories for each memory zone  
: 1 minute after last button operation  
: Temperature: 5 °C ~ 45 °C (41 °F~113 °F);  
Humidity: 40 ~ 85%  
Operation Environment  
Storage Environment  
: Temperature: -10 °C ~ 60 °C (14 °F~140 °F ); Humidity: 10 ~  
95%  
Power Source  
Dimensions  
Weight  
: 6V DC four LR06 (AA) Batteries  
: 137 (L) X 140 (W) X 63 (H) mm  
: 450g (G.W.) (No Batteries)  
Arm circumference  
: Adult: 24 ~36 cm (9.4”~14.2”) (standard)  
: Large adult: 34 ~46 cm (13.4”~18.1”) (standard)  
: Short time operation 2 minutes.  
K.B.  
*Specifications are subject to change without notice.  
16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Two-Year Limited Warranty:  
Lumiscope guarantees this product free from defects in material and workmanship for a period  
of two years from the date of purchase, except as noted below:  
This Lumiscope product warranty does not cover damage caused by misuse or abuse; the  
attachment of any unauthorized accessory; alteration to the product; or any other conditions  
whatsoever that are beyond the control of Lumiscope. Lumiscope shall not be responsible for  
any type of incidental, consequential, or special damage. All implied warranties, including, but  
not limited to those implied warranties of fitness and merchantability, are limited in the total  
duration of two years from the original date of purchase.  
To obtain warranty service on your Lumiscope Blood Pressure Monitor, please contact the  
fee for return shipping and handling. Please make checks payable to The Lumiscope Co., Inc.  
Upon receipt, Lumiscope will repair or replace, as appropriate, the blood pressure monitor and  
return it to you. Warranty is solely through the Lumiscope Repair Department. Service of this  
product by anyone other than the Lumiscope Repair Department voids warranty.  
In the event that Lumiscope does not abide by the terms of this warranty the consumer may  
seek breach of contract remedies in the New York State Federal courts of law. This warranty  
gives you specific legal rights and you may also have other rights which vary from state to state.  
Some states do not allow the exclusion or limitation of incidental or consequential damages, so  
the above limitation may not apply to you.  
Distributed By:  
The Lumiscope Co., Inc.  
Repair Department: 1-800-672-8293  
Made and Printed in China  
17  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Kuhn Rikon Mixer 5000 Series User Manual
Lexar Media MP3 Player LDP 600 User Manual
Life Fitness Home Gym SOBWS User Manual
Lucent Technologies Network Router 7820 0657 001 User Manual
Manitowoc Ice Ice Maker QM30 User Manual
Mark Levinson CD Player N390S User Manual
Master Appliance Welder UT 35 User Manual
Maytag Dishwasher ADB1500AWW User Manual
Maytag Washer Accessories 12001908 User Manual
MGE UPS Systems Power Supply EX15Rack User Manual