Invacare Camping Equipment MA50 User Manual

Owner’s Operator and Maintenance Manual  
MA50 Series  
MA50 Alternating Pressure  
Overlay System  
MA51 Alternating Pressure  
Mattress System  
MA55 Alternating Pressure  
Low Air Loss Mattress System  
DEALER: This manual MUST be given to  
the user of the product.  
USER: BEFORE using this product, read this  
manual and save for future reference.  
For more information regarding  
Invacare products, parts, and services,  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
TABLE OF CONTENTS  
TABLE OF CONTENTS  
Part No 1148136  
3
MA50 Series  
Download from Www.Somanuals.com. All Manuals Search And Download.  
TABLE OF CONTENTS  
TABLE OF CONTENTS  
REGISTER YOUR PRODUCT  
The benefits of registering include:  
1. Safeguarding your investment.  
2. Ensuring long-term maintenance and servicing of your product.  
3. Receiving updates with product information, maintenance tips and industry news.  
Register ONLINE at warranty.invacare.com  
Please have your model number and purchase date available to complete your registration.  
Any registration information you submit will only be used by Invacare Corporation and protected  
as required by applicable laws and regulations.  
MA50 Series  
4
Part No 1148136  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
TYPICAL PRODUCT PARAMETERS  
TYPICAL PRODUCT PARAMETERS  
Electrical Parameters  
MA50/MA51/MA55  
INPUT VOLTAGE AC:  
INPUT FREQUENCY:  
CURRENT:  
90 V  
60 Hz  
1 A  
5 ± 3 W  
MAXIMUM POWER CONSUMPTION:  
CIRCUIT PROTECTION:  
MODE OF OPERATION:  
Dual fused, 250 V, 1 A fast blow fuses  
Continuous  
Performance Parameters  
MA50/MA51/MA55  
WEIGHT CAPACITY  
STANDARD MATTRESS:  
BARIATRIC MATTRESS:  
PRESSURE ZONE:  
MAXIMUM FLOW:  
MAXIMUM FLOW PRESSURE:  
MAXIMUM FLOW TIMER:  
SUPPORT SURFACE INFLATION TIME:  
350 lbs  
1000 lbs  
2
8 ± 4 LPM  
35 ± 5 mmHg  
45 minutes  
20 - 45 minutes  
PATIENT COMFORT CONTROL PRESSURES  
SOFT PRESSURE:  
FIRM PRESSURE:  
8 ± 4 mmHg  
32 ± 4 mmHg  
5, 10, 15, 20 minutes  
CYCLE TIME:  
PATIENT CONTACT:  
Control unit and mattress have Latex-Free  
components  
Mechanical Parameters  
CONTROL UNIT  
MA50  
MA51  
MA55  
DIMENSIONS  
(L X W X H):  
9” x 5” x 5”  
WEIGHT:  
5 lbs  
POWER CORD:  
CONNECTION:  
PACKAGING:  
AIR FILTER:  
10 - 14 Feet Long, Hospital Grade  
Two ¼” Flow Couplings  
1 piece/box  
None  
Three ¼” Flow Couplings  
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
TYPICAL PRODUCT PARAMETERS  
Environmental Parameters  
MA50/MA51/MA55  
OPERATING CONDITIONS  
AMBIENT TEMPERATURE:  
RELATIVE HUMIDITY:  
50° - 104° F  
30% - 75% Non-Condensing  
70 - 106 kPa  
ATMOSPHERIC PRESSURE:  
STORAGE AND SHIPPING CONDITIONS  
AMBIENT TEMPERATURE:  
RELATIVE HUMIDITY:  
-40° - 158° F  
10% - 100% Non-Condensing  
50 - 106 kPa  
ATMOSPHERIC PRESSURE:  
6
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
LABEL LOCATION  
LABEL LOCATION  
DANGER-EXPLOSION HAZARD: DO NOT use in the presence of  
flammable anesthetics.  
CAUTION: Equipment should be connected to a properly grounded  
receptacle (3-prong). Risk of Electrical shock. DO NOT remove  
back. Disconnect air hose before administering CPR.  
NOTE: This label is on the back of  
the unit and also contains the  
serial number.  
WARNING  
Patient entrapment with bed side rails may cause injury or death. Mattress  
MUST fit bed frame and side rails snugly to prevent patient entrapment.  
Follow the manufacturer’s instructions. Monitor patient frequently. Read  
and understand the Owner’s/Operator’s Manual prior to using this  
or your dealer.  
Label part number 1150708 Rev A  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
SECTION 1—GENERAL GUIDELINES  
SECTION 1—GENERAL GUIDELINES  
WARNING  
SECTION 1 - GENERAL GUIDELINES contains important information for the safe  
operation and use of this product.  
Contraindications  
ALWAYS consult the patients physician before using the MA50, MA51 and MA55 systems.  
Installation  
The MA50, MA51 and MA55 systems MUST be installed on medical bed frames with side  
rails. The side rails MUST be in the raised position whenever a patient is on the bed.  
Controls on the footboard may be obstructed by the power unit on a few bed frames. It may  
be necessary to relocate the power unit. Refer to Installing the Power Unit on page 12.  
Check that air hoses and power cord are clear of moving bed components before placing a  
patient on the bed. Operate all bed frame motorized functions through their full range of  
motion to be certain that there is no pulling, interference or pinching.  
MA51 AND MA55 ONLY - DO NOT strap the mattress to the bed frame at the head and  
foot ends. Secure ALL mattress straps. Secure the straps to the bed deck at the head and foot  
ends and to the frame at the center of the bed. Otherwise damage to the mattress will occur  
when the head and foot ends are raised. Refer to Installing the Mattress on page 10.  
Entrapment May Occur  
WARNING  
Patient entrapment with bed side rails may cause injury or death. Mattress MUST  
fit bed frame and side rails snugly to prevent patient entrapment. Follow the  
manufacturer’s instructions. Monitor patient frequently. Read and understand the  
Owner’s/Operator’s Manual prior to using this equipment. Invacare product  
Proper patient assessment and monitoring, and proper maintenance and use of  
equipment is required to reduce the risk of entrapment. Variations in bed rail  
dimensions, and mattress thickness, size or density could increase the risk of  
entrapment. Review “A Guide to Bed Safety”, published by the Hospital Bed Safety  
product entry to access this bed safety guide.  
Refer to the owner’s manuals for beds and rails for additional product and safety  
information.  
After any adjustments, repair or service and before use, make sure all attaching  
hardware is tightened securely. Assist rails with dimensions different from the  
original equipment supplied or specified by the bed manufacturer may not be  
interchangeable and may result in entrapment or other injury.  
8
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
SECTION 1—GENERAL GUIDELINES  
Fire Hazard  
DANGER  
Smoking  
DO NOT SMOKE while using this device. This system uses room air for circulation  
through the mattress. A cigarette can burn a hole in the bed surface and cause  
damage to the mattress. Also, patient clothing, bed sheets, etc. may be combustible  
and cause a fire. Failure to observe this warning can result in severe fire, property  
damage and cause physical injury or death.  
Smoking by visitors in the room will contaminate the system. Therefore, visitor  
smoking is NOT permitted.  
Anesthesia Equipment  
There is an explosion risk if used with flammable anesthetics.  
Oxygen  
There is a possible fire hazard when used with oxygen administering equipment  
other than nasal mask or half bed tent type. The oxygen tent should NOT extend  
below mattress support level.  
Electrical  
DANGER  
Electrical shock hazard. DO NOT remove cover. Refer to qualified service  
personnel.  
Before performing any maintenance to the power unit, disconnect the power cord from the  
wall outlet.  
DO NOT insert items into any openings of the control unit. Doing so may cause fire or  
electric shock by shorting the internal components.  
The control unit MUST be kept away from all heat sources and radiators during operation.  
Connect the equipment to properly grounded three prong wall outlet using 10-14 ft hospital  
grade power cord provided with the product.  
Grounding reliability depends upon a properly grounded receptacle (3-prong).  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
SECTION 2—OPERATION  
SECTION 2—OPERATION  
CAUTION  
The control unit and mattress on the MA50 series are designed to be used as a  
system. DO NOT replace mattresses or control units with other models or other  
brands. Otherwise, damage to the system may occur. Contact your supplier to get  
the correct replacement if needed.  
Installing the Mattress  
Overlay System (MA50)  
NOTE: For this procedure, refer to FIGURE 2.1.  
NOTE: The overlay system comes with the following straps - head end elastic strap, foot end elastic  
strap and two buckle straps.  
1. If necessary, lower the side rails to facilitate installation of the mattress.  
2. Place the mattress overlay directly over the existing 3 to 5-inch foam mattress.  
3. Loop the head end elastic strap around the head end of the existing foam mattress.  
4. Loop the foot end elastic strap around the foot end of the existing foam mattress.  
5. Secure the overlay mattress to the existing foam mattress using the two buckle straps.  
NOTE: Ensure that the side rails can move freely.  
Mattress  
Overlay  
Head End  
Elastic Strap  
Foam  
Mattress  
Foot End  
Elastic Straps  
10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
SECTION 2—OPERATION  
Mattress Replacement System (MA51 and MA55)  
CAUTION  
DO NOT strap the mattress to the bed frame at the head and foot ends. Secure  
ALL mattress straps. Secure the straps to the bed deck at the head and foot ends  
and to the frame at the center of the bed. Otherwise damage to the mattress will  
occur when the head and foot ends are raised.  
Head End Bed Deck  
Center of Bed Frame  
Foot End Bed Deck  
DO NOT Strap Here  
NOTE: For this procedure, refer to FIGURE 2.2.  
NOTE: The powered mattress comes with ten nylon buckle straps.  
1. Remove the original foam mattress from the bed.  
2. If necessary, lower the side rails to facilitate installation of the mattress.  
3. Unroll the powered mattress and place it on the bed frame.  
NOTE: Ensure that the hose is towards the foot end of the bed.  
4. Use the buckle straps to secure the powered mattress to the bed deck in the following  
locations:  
• Head End - Head End Bed Deck  
• Foot End - Foot End Bed Deck  
• Center - Center of the Bed Frame  
Head End  
Powered Mattress  
Buckle  
Straps  
Hose  
Foot End  
11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
SECTION 2—OPERATION  
Installing the Side Rails  
WARNING  
Patient entrapment with bed side rails may cause injury or death. Mattress MUST  
fit bed frame and side rails snugly to prevent patient entrapment. Follow the  
manufacturer’s instructions. Monitor patient frequently. Read and understand the  
Owner’s/Operator’s Manual prior to using this equipment. Invacare product  
NOTE: Refer to the instructions provided with the side rails for the installation procedure.  
Installing the Power Unit  
NOTE: For this procedure, refer to  
Bed Hooks  
1. Swing out the bed hooks on the back of  
the control unit.  
2. Place the control unit on the footboard.  
NOTE: If the bed does not have a footboard,  
place the control unit on a flat surface, leaving  
room for the hose to hang down.  
Connecting the Hose  
CAUTION  
Ensure that the hose connecting the control unit to the mattress is routed such that  
it cannot be stepped on, kinked, squeezed or otherwise damaged.  
NOTE: For this procedure, refer to FIGURE 2.4 on page 13.  
1. Locate the hose at the foot end of the mattress.  
2. Locate the control unit connectors on the right side of the control unit.  
3. Squeeze and hold the tabs on the hose connectors.  
4. Push the hose connectors onto the control connectors until an audible click is heard.  
NOTE: The audible click indicates that the hose connectors are properly engaged with the control  
connectors.  
MA50 Series  
12  
Part No 1148136  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
SECTION 2—OPERATION  
Control Unit Connectors  
Tab  
Hose  
Connector  
Hose  
Connecting the Power Cord  
WARNING  
DO NOT alter plug to fit a non-conforming outlet. Instead, have an electrician  
install a properly grounded 3-prong outlet. Failure to use the correct plug and outlet  
can result in a potential safety hazard.  
CAUTION  
Ensure that the power cord of the control unit is not pinched, or has any objects  
placed on it, and also ensure it is not located where it can be stepped on or tripped  
over.  
1. Examine the hospital grade power cord supplied with the control unit.  
2. Perform one of the following:  
• If the plug is damaged - Call your supplier for a replacement hospital grade cord.  
• If the plug is not damaged - Plug the end of the supplied hospital grade power cord  
into the power outlet on the side of the control unit.  
3. Plug the other end of the plug into a properly grounded outlet on the wall.  
NOTE: Once the unit is plugged in, an AMBER LED on the control unit is lit indicating that the  
system is in STAND BY mode.  
Part No 1148136  
13  
MA50 Series  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
SECTION 2—OPERATION  
Using the Front Panel  
NOTE: For this procedure, refer to FIGURE 2.5.  
DETAIL “A” - MA50 AND MA51 FRONT PANEL  
Power Fail LED  
Low Pressure  
LED  
Max Inflate  
Button  
Fowler  
Button  
Firm Button  
Lock Button  
Soft Button  
Alarm Silence  
Button  
Static/Alternating  
Pressure Button  
Power  
Button  
LCD  
Display  
DETAIL “B” - MA55 FRONT PANEL  
Power Fail LED  
Low Pressure  
LED  
Mode Button  
(Max Inflate/Low  
Air Loss)  
Fowler  
Button  
Soft Button  
Firm Button  
Alarm Silence  
Button  
LCD  
Display  
Static/Alternating  
Pressure Button  
Power  
Button  
Lock Button  
Power Button  
1. To turn the control unit on or off, press and release the Power button (  
).  
NOTE: Once the unit is plugged in, an AMBER LED on the control unit is lit indicating that the  
system is in STAND BY mode. Once the Power button is pressed and released, a GREEN LED  
illuminates indicating that the control unit is on.  
NOTE: If the power comes on after a power outage, the control unit will go through its system  
initialization routine for a few seconds and then resume the desired function.  
MA50 Series  
14  
Part No 1148136  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
SECTION 2—OPERATION  
Static/Alternating Pressure Button  
1. Press the Alternating Pressure button (  
of four Alternating Pressure Times.  
) to select between Static Mode and one  
NOTE: The alternating pressure times indicate the frequency of deflation or inflation of half of the  
air cushions (even or odd numbered). For example, 10 minutes is selected. In this example, the even  
numbered air cushions in the mattress will deflate, while the odd numbered cushions remain at  
constant pressure. When 10 minutes has elapsed, the even numbered air cushions will inflate while  
the odd numbered cushions remain at constant pressure. When 10 minutes has elapsed again, the  
odd numbered air cushions deflate while the even numbered cushions remain at constant pressure.  
The air cushions continue to inflate or deflate like this as long as an alternating pressure time has  
been selected. At all times, half of the air cushions remain at a constant pressure when an  
alternating pressure time has been selected.  
NOTE: In Static Mode, all air cushions in the mattress are maintained at constant pressure. The  
Static LED is illuminated GREEN.  
Adjusting the Minimum Pressure Setting (A/P Low)  
The minimum pressure setting during Alternating Pressure is set to 0 (2 mmHg) at the  
factory. This means that the mattress will deflate as much as possible during the deflation  
time when using 10, 15 or 20 minute Alternating Pressure times. The setting can be  
changed so the mattress will only deflate to 50% of the selected pressure setting.  
1. Ensure the Standby LED is lit.  
2. Perform one of the following:  
• MA50 and MA51 - Press and hold the Max Inflate and Fowler buttons.  
• MA55 - Press and hold the Mode and Fowler buttons.  
3. Read the LCD Display. The LCD Display will flash:  
• 0 if the current setting is 0 (2 mmHg).  
• 5 if the current setting is 50% of the selected pressure.  
4. Perform one of the following to toggle between the two settings:  
• MA50 and MA51 - Press the Static/Alternating Pressure button.  
• MA55 - Press the Static/Alternating Pressure button.  
5. Press the Power button to save the setting.  
Part No 1148136  
15  
MA50 Series  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
SECTION 2—OPERATION  
Firm/Soft Buttons  
1. Select comfort pressure settings by pressing Firm (  
) or Soft (  
) button.  
• Soft Button - Pressing this button reduces the pressure setting in the mattress.  
• Firm Button - Pressing this button increase the pressure setting in the mattress.  
NOTE: The patient comfort pressure ranges from Soft (level 0 = 6±4 mmHg) to Firm (level 9 =  
32±6 mmHg). The Comfort Control LED displays the patient comfort pressure levels from 0 to 9  
and provides a guide to the caregiver to set approximate comfort pressure level depending on the  
patient weight. If the patient’s weight to height ratio is above average, increase the pressure setting  
by approximately 20%.  
Fowler Button  
1. Press the Fowler button (  
) to activate the patient fowler mode.  
NOTE: The Fowler LED illuminates when in this mode. When this mode is activated, the control  
unit increases the pressure in the mattress to prevent the patient from bottoming out.  
Max Inflate Button  
NOTE: This button is on the MA50 and MA51 only.  
1. Press the Max Inflate button (  
) to select the Max Inflate mode and inflate the  
mattress rapidly to maximum firmness (pressurized to 35 ± 6 mmHg). A series of  
beeps sound every three minutes as a reminder that the Max Inflate mode is active.  
After 30 minutes, the Max Inflate mode deactivates and the control unit defaults to the  
previous setting.  
NOTE: It takes 20-45 minutes for the mattress to inflate fully (inflation time depends on size of  
mattress).  
NOTE: Max Inflate mode can be manually disengaged by pressing the Max Inflate button. This  
will deactivate the Max Inflate LED.  
NOTE: It is recommended that Max Inflate setting be used during patient ingress/egress, patient  
wound care, patient turning or patient cleaning.  
2. Press the Max Inflate button again to manually disengage Max Inflate mode.  
NOTE: This will deactivate the Max Inflate LED.  
MA50 Series  
16  
Part No 1148136  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
SECTION 2—OPERATION  
Mode Button (Max Inflate/Low Air Loss)  
NOTE: The Mode button is on the MA55 system only.  
1. Press the Mode button (  
mode.  
) to select the Max Inflate mode or the Low Air Loss  
• Max Inflate Mode - In this mode, the mattress inflates rapidly to maximum  
firmness (pressurized to 35 ± 6 mmHg). A series of beeps sound every three  
minutes as a reminder that the Max Inflate mode is active. After 45 minutes, the  
Max Inflate mode deactivates and the control unit defaults to the previous setting.  
NOTE: It takes 20-45 minutes for the mattress to inflate fully (inflation time depends on size  
of mattress).  
NOTE: It is recommended that Max Inflate setting be used during patient ingress/egress,  
patient wound care, patient turning or patient cleaning.  
• Low Air Loss Mode - In this mode, the mattress goes into on-demand low air loss  
relief mode.  
2. Press the Mode button to manually disengage Max Inflate or Low Air Loss mode.  
NOTE: This will deactivate the Max Inflate and Low Air Loss LEDs.  
Lock Button  
1. Press the Lock button (  
) to lock out all control unit functions, including the On  
button, to prevent any tampering with the settings. Press the button for approximately  
3-5 seconds for the Lock LED to activate.  
NOTE: The Lock LED illuminates when in this mode.  
Alarm Silence Button  
1. Press the Alarm Silence button ( ) to silence the alarm that sounds in the event  
of power failure or when the hose is disconnected from the control unit.  
NOTE: The Alarm LED illuminates when in this mode.  
Power Fail LED  
In the event of power outage, an alarm sounds and Power Fail LED flashes AMBER. The  
control unit has internal memory and retains the previous settings during the power  
outage.  
NOTE: During a power outage, the mattress retains the air as long as the mattress is connected to  
the control unit.  
Part No 1148136  
17  
MA50 Series  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
SECTION 2—OPERATION  
Low Pressure LED  
In the event that the mattress hose disconnects, an alarm sounds and Low Pressure LED  
flashes AMBER. Once the low pressure problem is fixed, the control unit resumes  
operation in the previously set mode.  
Powering Up the System  
NOTE: For this procedure, refer to FIGURE 2.5 on page 14.  
1. Turn on the power to the system by pressing the Power button on the control unit.  
NOTE: Once the button is released, a GREEN LED illuminates when the unit is on.  
Placing the Patient on the Mattress  
1. Perform one of the following:  
• MA50 or MA51 - Press the Max Inflate button to turn on the control unit to  
maximum flow.  
• MA55 - Press the Mode button to select Max Inflate Mode.  
NOTE: In this mode, the Max Inflate LED lights up.  
2. After the mattress is fully inflated, place the patient on the mattress.  
3. Ensure that the patients feet are towards the foot end of the mattress (the end with the  
hose).  
4. Center the patient on the bed from side-to-side and head-to-foot.  
NOTE: Special positioning may be required with contracted patients to provide comfortable  
positions.  
5. After placing the patient, make certain no objects will fall under the patient, such as  
feeding tubes, IV's etc.  
6. Wait five minutes for the mattress pressure to stabilize.  
7. Once the mattress inflates to its normal size, set the comfort pressure to the desired  
comfort level.  
8. Wait five minutes for the mattress pressure to stabilize.  
MA50 Series  
18  
Part No 1148136  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
SECTION 2—OPERATION  
9. Verify that the patient has not bottomed out by performing the following steps:  
A. Ensure that the patient is lying flat on his/her back in the middle of the mattress.  
B. Place four fingers between the air cushions directly underneath the sacral region  
of the patients body.  
C. Ensure that there is 3 to 4-finger width clearance between the bottom of the patient  
and the bed frame.  
D. Adjust the comfort setting, if needed.  
E. Wait five minutes for the mattress pressure to stabilize.  
F. Repeat STEPS A-E until patient has not bottomed out and patient comfort is  
achieved.  
10. If the patient feels that the bed is too soft/hard, press the Soft or Firm button to adjust  
the comfort settings.  
11. Use a regular pillow to help support and stabilize the patient's head.  
Transferring Patient From/To a Gurney  
WARNING  
ALWAYS engage the wheel locks of the bed and the wheel locks of the gurney  
before transferring the patient between the bed and the gurney.  
1. Engage the wheel locks of the bed. Refer to the owners manual provided with the bed.  
2. Engage the wheel locks of the gurney. Refer to the owners manual provided with the  
gurney.  
3. Press the Max Inflate or Mode button to achieve maximum mattress pressure.  
4. Raise or lower the bed to match the gurney height. Refer to the owners manual  
provided with the bed.  
5. When the mattress has reached maximum firmness, perform one of the following:  
• Bed to Gurney Transfer - Slide the patient onto the gurney.  
• Gurney to Bed Transfer - Slide the patient onto the bed.  
Part No 1148136  
19  
MA50 Series  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
SECTION 2—OPERATION  
Transferring Patient From/To a Wheelchair  
WARNING  
ALWAYS engage the wheel locks of the bed and the wheel locks of the wheelchair  
before transferring the patient between the bed and the wheelchair.  
1. Engage the wheel locks of the bed. Refer to the owners manual provided with the bed.  
2. Engage the wheel locks of the wheelchair, if applicable. Refer to the owners manual  
provided with the wheelchair.  
3. Press the Max Inflate or Mode button to achieve maximum mattress pressure.  
4. Raise or lower the bed to match the wheelchair height. Refer to the owners manual  
provided with the bed.  
5. When the mattress has reached maximum firmness, perform one of the following:  
• Bed to Wheelchair Transfer - Slide the patient onto the wheelchair.  
• Wheelchair to Bed Transfer - Slide the patient onto the bed.  
Preparing for CPR Procedure  
NOTE: For this procedure, refer to FIGURE 2.6.  
1. Press and hold the tabs on hose connector while pulling the hose from the control unit.  
2. Disconnect the RED CPR connector located on the side of the mattress.  
Control Unit Connectors  
Mattress  
Tab  
Hose  
RED CPR  
Connector  
Connector  
Hose  
About Power Outage and Transportation  
If the hoses remain connected to the control unit, the mattress retains air during a power  
outage, during transportation, when the control unit is unplugged or when the control  
unit is turned Off. If the hose becomes disconnected or damaged and the mattress  
deflates, the mattress has a 2-inch foam pad to provide patient support .  
MA50 Series  
20  
Part No 1148136  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
SECTION 3—MAINTENANCE AND TROUBLESHOOTING  
SECTION 3—MAINTENANCE AND  
TROUBLESHOOTING  
Cleaning the System  
WARNING  
Before cleaning or disassembling the MA50 series, check the underside of the  
mattress folds for sharp objects such as scissors, needles, etc. These objects should  
be removed and discarded before proceeding with further cleaning or disassembly;  
otherwise injury or damage to the product may occur.  
Because of potential risk of infectious exposure, cleaning with the patient on the bed  
is not recommended.  
All equipment should be inspected. Any item that is visibly soiled with the patient's  
blood or other body fluids should be properly cleaned or removed.  
Laundry workers should treat all soiled bedding as if it were contaminated with  
pathogenic microorganisms.  
1. Remove the bedding.  
2. If necessary, inflate the mattress.  
3. Ensure that the control unit is off.  
4. Unplug the power cord from the wall outlet.  
5. Ensure that the underside of the mattress is clear of all sharp objects.  
6. Examine the surface of the control unit and mattress assembly components for visible  
blood or body fluids.  
7. Perform one of the following:  
• If blood is present, decontaminate the product.  
i. Remove all visible soil with disposable paper towels.  
ii. Scrub the area with freshly prepared effective phenolic detergent disinfectant  
solution.  
• If blood is not present, remove any soil from the cover with paper towels.  
NOTE: If soiled, the cover should be removed, cleaned and decontaminated.  
8. Perform the following steps to clean the mattress cover:  
A. Using a clean sponge or paper towel, wipe down the cover surface with a dilute  
detergent solution of quaternary cleaner disinfectant or other germicidal detergent  
solution.  
B. Remove the cover and launder it using the following method:  
21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
SECTION 3—MAINTENANCE AND TROUBLESHOOTING  
CAUTION  
Laundry workers should always wash their hands before working with clean  
bedding.  
DO NOT overload the machine.  
DO NOT use chlorine bleach because it may damage the fabric coating.  
High air temperatures will damage the fabrics and void the Invacare warranty.  
i. Place the cover in a washing machine.  
ii. Wash with warm water (below 120°F).  
iii. Add detergent and disinfectant according to the manufacturers instructions.  
iv. Remove excess water.  
v. Set the dryer to the lowest setting (below 120°F).  
vi. Dry the cover until it is completely dry.  
9. Perform the following steps to clean the control unit and hose fittings:  
A. Wipe all controls, chassis and hose fittings with a quaternary disinfectant solution.  
B. Using a nylon brush, gently clean all crevices as they can harbor microorganisms.  
C. Air dry all treated surfaces.  
10. Perform the following steps to clean the mattress components (air cushions, mattress  
base, etc.):  
A. Using a clean sponge or paper towel, wipe down the mattress components with a  
dilute detergent solution of quaternary cleaner disinfectant or other germicidal  
detergent solution.  
B. Wipe the mattress components with a clean dry cloth.  
Storing the System  
1. Ensure that the control unit is off and disconnect the power cord from the wall outlet.  
3. Disconnect the air hose connector(s) from the control unit and allow air to vent from  
the mattress.  
4. Gently roll up the mattress with minimal handling and agitation.  
5. Ensure the cover surface is inside the roll.  
6. Store the unit, keeping the mattress with the control unit.  
22  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
SECTION 3—MAINTENANCE AND TROUBLESHOOTING  
Troubleshooting  
PROBLEM  
CAUSE  
SOLUTION  
Mattress not inflating  
Not alternating properly  
Mattress hose disconnected  
Connect hose connectors and lock  
them in place  
Air hose kinked or split  
Unkink hose or replace split hose  
Major leak in the air cushions or overlay pad Replace leaking air cushions or  
overlay pad  
Kinked or split manifold  
Unkink manifold or replace split  
manifold  
Has power and fuse is good, control unit  
does not come on  
Send control unit back to factory for  
repair  
Not alternating, solenoid malfunction  
Send control unit back to factory for  
repair  
No air, pump malfunction  
Control unit off  
Send unit for repair  
No power  
Check power source and turn on unit  
Power cord disconnected  
Connect power cord to the power  
source  
No power in the power source  
Check power source has power and  
turn it On  
Power outage  
Blown fuse  
Wait till the power source has power  
Replace blown fuse with an equivalent  
fuse  
Part No 1148136  
23  
MA50 Series  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
LIMITED WARRANTY  
PLEASE NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW  
APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4, 1975.  
This warranty is extended only to the original purchaser who purchases this product when new and unused  
from Invacare or a dealer. This warranty is not extended to any other person or entity and is not transferable  
or assignable to any subsequent purchaser or owner. Coverage under this warranty will end upon any such  
subsequent sale or other transfer of title to any other person.  
This warranty gives you specific legal rights and you may also have other legal rights which vary from state to  
state.  
Invacare warrants the mattress, mattress overlay and cover when purchased new and unused to be free from  
defects in materials and workmanship for a period of one year from the date of purchase from Invacare or a  
dealer, with a copy of the seller’s invoice required for coverage under this warranty. Invacare warrants  
thecontrol unit when purchased new and unused to be free from defects in materials and workmanship for a  
period of two years from the date of purchase from Invacare or a dealer, with a copy of the seller’s invoice  
required for coverage under this warranty.If within such warranty period any such product shall be proven to  
be defective, such product shall be repaired or replaced, at Invacare's option. This warranty does not include  
any labor or shipping charges incurred in replacement part installation or repair of any such product. Invacare's  
sole obligation and your exclusive remedy under this warranty shall be limited to such repair and/or  
replacement.  
For warranty service, please contact the dealer from whom you purchased your Invacare product. In the event  
you do not receive satisfactory warranty service, please write directly to Invacare at the address on the back  
cover. Provide dealer's name, address, model number, and the date of purchase, indicate nature of the defect  
and, if the product is serialized, indicate the serial number.  
Invacare Corporation will issue a return authorization. The defective unit or parts must be returned for  
warranty inspection using the serial number, when applicable, as identification within thirty days of return  
authorization date. DO NOT return products to our factory without our prior consent. C.O.D. shipments will  
be refused; please prepay shipping charges.  
LIMITATIONS AND EXCLUSIONS: THE WARRANTY SHALL NOT APPLY TO PROBLEMS ARISING  
FROM NORMAL WEAR OR FAILURE TO ADHERE TO THE ENCLOSED INSTRUCTIONS. IN ADDITION,  
THE FOREGOING WARRANTY SHALL NOT APPLY TO SERIAL NUMBERED PRODUCTS IF THE  
SERIAL NUMBER HAS BEEN REMOVED OR DEFACED; PRODUCTS SUBJECTED TO NEGLIGENCE,  
ACCIDENT, IMPROPER OPERATION, MAINTENANCE OR STORAGE; OR PRODUCTS MODIFIED  
WITHOUT INVACARE'S EXPRESS WRITTEN CONSENT INCLUDING, BUT NOT LIMITED TO:  
MODIFICATION THROUGH THE USE OF UNAUTHORIZED PARTS OR ATTACHMENTS: PRODUCTS  
DAMAGED BY REASON OF REPAIRS MADE TO ANY COMPONENT WITHOUT THE SPECIFIC  
CONSENT OF INVACARE; PRODUCTS DAMAGED BY CIRCUMSTANCES BEYOND INVACARE'S  
CONTROL; PRODUCTS REPAIRED BY ANYONE OTHER THAN AN INVACARE DEALER, SUCH  
EVALUATION SHALL BE SOLELY DETERMINED BY INVACARE.  
THE FOREGOING EXPRESS WARRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER EXPRESS  
WARRANTIES WHATSOEVER, WHETHER EXPRESS OR IMPLIED, INCLUDING THE IMPLIED  
WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE AND THE SOLE  
REMEDY FOR VIOLATIONS OF ANY WARRANTY WHATSOEVER, SHALL BE LIMITED TO REPAIR OR  
REPLACEMENT OF THE DEFECTIVE PRODUCT PURSUANT TO THE TERMS CONTAINED HEREIN.  
THE APPLICATION OF ANY IMPLIED WARRANTY WHATSOEVER SHALL NOT EXTEND BEYOND  
THE DURATION OF THE EXPRESS WARRANTY PROVIDED HEREIN. INVACARE SHALL NOT BE  
LIABLE FOR ANY CONSEQUENTIAL OR INCIDENTAL DAMAGES WHATSOEVER.  
THIS WARRANTY SHALL BE EXTENDED TO COMPLY WITH STATE/PROVINCIAL LAWS AND  
REQUIREMENTS.  
Invacare Corporation  
USA  
Canada  
570 Matheson Blvd E All rights reserved. Trademarks are identified by  
Unit 8 the symbols ™ and ®. All trademarks are owned  
Mississauga Ontario by or licensed to Invacare Corporation unless  
One Invacare Way  
Elyria, Ohio USA  
44036-2125  
800-333-6900  
L4Z 4G4 Canada  
800-668-5324  
otherwise noted.  
© 2008 Invacare Corporation  
Part No 1148136  
Rev B - 02/08  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   

Indesit Washer IWSD 61251 User Manual
Ingersoll Rand Grinder 302A User Manual
Insignia Stereo System NS A3111 User Manual
Invacare Mobility Aid 1130200 User Manual
JVC DVR D VHS User Manual
JVC VCR HR A5U User Manual
Kenwood Car Stereo System 68 User Manual
Kenwood CD Player KDC Z737 User Manual
Kodak Camcorder z1015 User Manual
Kodak Camera Accessories K625 User Manual