Pharmaceutical Refrigerator
Operation Manual
HYC-260-360
Read the Operation Manual carefully before using your appliance.
Keep the Operation Manual in a safe place.
Appearance, color and layout of the door may vary.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
1) The appliance should be connected to a dedicated socket for 220V/50Hz power
supply.Allow some time lapse before loading the appliance.
Battery
switch
Lamp
switch
Control
pannel
2) The warning sound signal may be heard when the appliance is first connected to the
power source and can be inactivated by pressing the buzzer buttom. The warning
indicator stays on until the temperature of the test bottle sensor reaches 5
2
.
Inner light
3) Ensure that the two test bottles are filled with 10% glycol solution.(see article 4,section
installation on page 6).
4) The appliance is preset during manufacture for operation temperature 2~8
.
Door handle
Casters
Shelf
Download from Www.Somanuals.com. All Manuals Search And Download.
1.Control panel
The operation mode and temperature are displayed on the control panel. Pressing the
button on the control panel will display the current temperature and warning data.
2.Light switch
Control of the lights inside the appliance.
3.Battery switch
Rechargeable battery switch for control of power interruption. Position 1 for On, and
Position 0 for Off.
Caution: The switch must be set to position On while the appliance is in operation.
Caution: Set the battery switch to position Off if the appliance is not used for a long
period.
4.Door handle
Elegant design and unique finish.
5.Casters
Casters designed for free movement of the appliance.
6.Shelf
7.Inner light
15W
8
Download from Www.Somanuals.com. All Manuals Search And Download.
Temperature detected by
upper or lower sensor
Tempreture detected by
upper or lower sensor
8
2
Temperature detected by
sensor or
8
2
Download from Www.Somanuals.com. All Manuals Search And Download.
1.Temperature display
Preset temperature of the appliance is 5
2
.
Preset warning point of high temperature is 8 ,and of low temperature is 2
.
Caution:Displayed temperature is detected by test bottles for upper and lower sections of
the appliance, and does not always remain at 5 .The information displayed by the snesor
indicates the average temperature.
Temperature adjusting:
Press and hold both SENSOR SELECT button and CALIB CANCEL button for more than 3s until
original setting starts flashing to indicate that it enters temperature setting mode. With each press
of CALIB CANCEL button, temperature setting increases 0.1 until reaching upper limit, which
is defaulted as 8 . If press again, it returns to lower limit, which is defaulted as 2 . It repeats the
above cycle. When it reaches your desired temperature, if you don't press the button within 5s,
the temperature setting will be saved and it automatically quits the temperature setting mode (the
temperature setting range programmed at factory is 2 - 8 ). The display returns to display inside
sensor temperature or average temperature before setting is made (not flashing).
Download from Www.Somanuals.com. All Manuals Search And Download.
|