®
80 Series V
Digital Multimeter
Calibration Manual
PN 2102915
September 2004, Rev.1, 2/05
© 2004-2005 Fluke Corporation, All rights reserved. Printed in USA
All product names are trademarks of their respective companies.
Download from Www.Somanuals.com. All Manuals Search And Download.
Table of Contents
Title
Page
Introduction........................................................................................................ 1
Contacting Fluke................................................................................................ 1
Precautions and Safety Information................................................................... 2
Electrical Symbols ............................................................................................. 4
Specifications..................................................................................................... 4
General Specifications................................................................................... 4
Detailed Specifications.................................................................................. 5
Basic Maintenance............................................................................................. 9
Cleaning the Meter ........................................................................................ 9
Opening the Meter Case ................................................................................ 9
Reassembling the Meter Case........................................................................ 11
Replacing the Battery .................................................................................... 12
Testing Fuses and Current Circuitry.............................................................. 12
Replacing the Fuses....................................................................................... 13
Required Equipment .......................................................................................... 13
Performance Tests.............................................................................................. 14
Basic Operability Tests.................................................................................. 15
Testing the Fuses........................................................................................... 15
Testing the Display........................................................................................ 15
Testing the Pushbuttons................................................................................. 17
Testing Meter Accuracy ................................................................................ 17
Calibration Adjustment...................................................................................... 23
Calibration Adjustment Counter.................................................................... 23
Calibration Adjustment Password ................................................................. 23
Changing the Password............................................................................. 23
Calibration Adjustment Procedure ................................................................ 25
Service and Parts................................................................................................ 27
i
Download from Www.Somanuals.com. All Manuals Search And Download.
80 Series V
Calibration Manual
ii
Download from Www.Somanuals.com. All Manuals Search And Download.
List of Tables
Table
Title
Page
1.
2.
3.
4.
5.
6.
7.
8.
9.
Electrical Symbols ................................................................................................. 4
Temperature Specifications (87 and 88 Only) ....................................................... 6
Current Function Specifications............................................................................. 6
Frequency Counter Specifications (Model 88) ...................................................... 7
10. Frequency Counter Sensitivity and Trigger Levels................................................ 7
11. Electrical Characteristics of the Terminals............................................................. 8
12. MIN MAX Recording Specifications .................................................................... 8
13. Required Tools and Equipment.............................................................................. 14
14. Display Features..................................................................................................... 15
15. Accuracy Tests....................................................................................................... 17
16. Calibration Adjustment Steps................................................................................. 26
17. 80 Series V Final Assembly................................................................................... 27
iii
Download from Www.Somanuals.com. All Manuals Search And Download.
80 Series V
Calibration Manual
iv
Download from Www.Somanuals.com. All Manuals Search And Download.
List of Figures
Figure
Title
Page
1.
2.
3.
4.
5.
6.
7.
8.
Removing LCD Mask to Access LCD................................................................... 11
Testing the Current Input Fuses ............................................................................. 13
Display Features..................................................................................................... 15
Setup for Inductive Pickup Test............................................................................. 21
Waveform for Inductive Pickup Test..................................................................... 22
Restoring the Default Password ............................................................................ 24
80 Series V Final Assembly................................................................................... 28
v
Download from Www.Somanuals.com. All Manuals Search And Download.
80 Series V
Calibration Manual
vi
Download from Www.Somanuals.com. All Manuals Search And Download.
Introduction
XWWarning
To avoid shock or injury:
•
•
•
Read “Precautions and Safety Information” before
performing the verification tests or calibration adjustment
procedures documented in this manual.
Do not perform the verification tests or calibration
adjustment procedures described in this manual unless you
are qualified to do so.
The information provided in this manual is for the use of
qualified personnel only.
WCaution
•
•
The 80 Series V Digital Multimeters contains parts that can
be damaged by static discharge.
Follow the standard practices for handling static sensitive
devices.
The 80 Series V Calibration Manual provides the following information:
•
•
•
•
•
•
Safety information
Specifications
Basic maintenance (cleaning, replacing the battery and fuses)
Performance test procedures
Calibration adjustment procedures
Replaceable parts
For complete operating instructions, refer to the 80 Series V or 88 Series V Users
Manual.
Contacting Fluke
To contact Fluke or locate the nearest Service Center, call one of the following telephone
numbers:
USA: 1-888-44-FLUKE (1-888-443-5853)
Canada: 1-800-36-FLUKE (1-800-363-5853)
Europe: +31 402-675-200
Japan: +81-3-3434-0181
Singapore: +65-738-5655
Anywhere in the world: +1-425-446-5500
1
Download from Www.Somanuals.com. All Manuals Search And Download.
80 Series V
Calibration Manual
XWWarning
To avoid possible electric shock or personal injury, inspect the
test leads for damaged insulation or exposed metal. Check the
test leads for continuity. Replace damaged test leads before
using the Meter.
Precautions and Safety Information
In this manual, a Warning identifies conditions and actions that pose hazard(s) to the
user; a Caution identifies conditions and actions that may damage the Meter or the test
instruments.
XWWarning
To avoid possible electric shock or personal injury, follow these
guidelines:
•
Use this Meter only as specified in this manual or the
protection provided by the Meter might be impaired.
•
Do not use the Meter if it is damaged. Before using the
Meter, inspect the case. Look for cracks or missing plastic.
Pay particular attention to the insulation surrounding the
connectors.
•
•
•
•
Make sure the battery door is closed and latched before
operating the Meter.
Replace the battery as soon as the battery indicator (M)
appears.
Remove test leads from the Meter before opening the
battery door.
Inspect the test leads for damaged insulation or exposed
metal. Check the test leads for continuity. Replace damaged
test leads before using the Meter.
•
Do not apply more than the rated voltage, as marked on the
Meter, between the terminals or between any terminal and
earth ground.
•
•
Never operate the Meter with the cover removed or the case
open.
Use caution when working with voltages above 30 V ac rms,
42 V ac peak, or 60 V dc. These voltages pose a shock
hazard.
•
•
Use only the replacement fuses specified in this manual.
Use the proper terminals, function, and range for
measurements.
•
•
Avoid working alone.
When measuring current, turn off circuit power before
connecting the Meter in the circuit. Remember to place the
Meter in series with the circuit.
2
Download from Www.Somanuals.com. All Manuals Search And Download.
Digital Multimeter
Precautions and Safety Information
•
When making electrical connections, connect the common
test lead before connecting the live test lead; when
disconnecting, disconnect the live test lead before
disconnecting the common test lead.
•
•
•
•
•
•
Do not use the Meter if it operates abnormally. Protection
may be impaired. When in doubt, have the Meter serviced.
Do not operate the Meter around explosive gas, vapor, or
dust.
Use only a single 9 V battery, properly installed in the Meter
case, to power the Meter.
When servicing the Meter, use only specified replacement
parts.
When using probes, keep fingers behind the finger guards
on the probes.
Do not use the Low Pass Filter option to verify the presence
of hazardous voltages. Voltages greater than what is
indicated may be present. Make a voltage measurement
without the filter to detect the possible presence of
hazardous voltage, then select the filter function.
XCaution
To avoid possible damage to the Meter or to the equipment
under test, follow these guidelines:
•
Disconnect circuit power and discharge all high-voltage
capacitors before testing resistance, continuity, diodes, or
capacitance.
•
Before measuring current, check the Meter's fuses.
See " Testing the Fuses”.
3
Download from Www.Somanuals.com. All Manuals Search And Download.
80 Series V
Calibration Manual
Electrical Symbols
Electrical symbols used on the Meter and in this manual are explained in Table 1.
Table 1. Electrical Symbols
AC (Alternating Current)
DC (Direct Current)
Hazardous voltage.
Earth ground
B
J
Fuse
F
I
P
Conforms to European Union directives
X
Risk of Danger. Important
information. See Manual.
Conforms to relevant Canadian
Standards Association directives
W
M
R
$
T
Battery
Double insulated
Capacitance
Continuity test or continuity beeper
tone.
E
G
Underwriters Laboratories
IEC overvoltage category III
Diode
t
IEC overvoltage category IV
CAT III equipment is designed to
protect against transients in
equipment in fixed-equipment
installations, such as distribution
panels, feeders and short branch
circuits, and lighting systems in
large buildings.
CAT IV equipment is designed to protect
against transients from the primary
supply level, such as an electricity meter
or an overhead or underground utility
service.
CAT III
CAT IV
Inspected and licensed by TÜV Product Services.
s
Specifications
General Specifications
Maximum Voltage between any Terminal and Earth Ground: 1000 V rms
WFuse Protection for mA or µA inputs: 44/100 A, 1000 V FAST Fuse
WFuse Protection for A input: 11 A, 1000 V FAST Fuse
Display: Digital: 6000 counts updates 4/sec; (Model 87 and 88 also has 19,999 counts in high-resolution mode).
Analog Bargraph: 33 segments, updates 40/sec. Frequency: 19,999 counts, updates 3/sec at > 10 Hz
Temperature: Operating: -20 °C to +55 °C; Storage: -40 °C to +60 °C
Altitude: Operating: 2000 m; Storage: 10,000 m
Temperature Coefficient: 0.05 x (specified accuracy)/ °C (< 18 °C or > 28 °C)
Electromagnetic Compatibility: In an RF field of 3 V/m total accuracy = specified accuracy + 20 counts
Except: 600 µA dc range total accuracy=specified accuracy + 60 counts.
Temperature not specified.
Relative Humidity: 0 % to 90 % (0 °C to 35 °C); 0 % to 7 0% (35 °C to 55 °C)
Battery Type: 9 V zinc, NEDA 1604 or 6F22 or 006P
Battery Life: 400 hrs typical with alkaline battery (with backlight off)
Vibration: Per MIL-PRF-28800 for a Class 2 instrument
Shock: 1 Meter drop per IEC 61010-1:2001
Size (HxWxL): 1.25 in x 3.41 in x 7.35 in (3.1 cm x 8.6 cm x 18.6 cm)
Size with Holster and Flex-Stand: 2.06 in x 3.86 in x 7.93 in (5.2 cm x 9.8 cm x 20.1 cm)
Weight: 12.5 oz (355 g)
Weight with Holster and Flex-Stand: 22.0 oz (624 g)
Safety: Complies with ANSI/ISA S82.01-2004, CSA 22.2 No. 1010.1:2004 to 1000 V Overvoltage Category III, IEC
4
Download from Www.Somanuals.com. All Manuals Search And Download.
Digital Multimeter
Specifications
664 to 600 V Overvoltage Category IV. UL listed to UL61010-1. Licensed by TÜV to EN61010-1.
Detailed Specifications
For all detailed specifications:
Accuracy is given as ±([% of reading] + [number of least significant digits]) at 18° C to
28° C, with relative humidity up to 90 %, for a period of one year after calibration.
For Model 87 in the 4 ½-digit mode, multiply the number of least significant digits
(counts) by 10. AC conversions are ac-coupled and valid from 3 % to 100 % of range.
Model 87 is true rms responding. AC crest factor can be up to 3 at full scale, 6 at half
scale. For non-sinusoidal wave forms add -(2 % Rdg + 2 % full scale) typical, for a crest
factor up to 3.
Table 2. Model 87 AC Voltage Function Specifications
Function
Range
Resolution
Accuracy
440 Hz -
1 kHz
1
45 - 65 Hz 30 - 200 Hz 200 - 440 Hz
1 - 5 kHz
5 - 20 kHz
600.0 mV
6.000 V
60.00 V
600.0 V
0.1 mV
0.001 V
0.01 V
0.1 V
± (0.7 % + 4)
± (2.0 % + 4)
± (2.0 % + 20)
K2,4
± (1.0 % + 4)
± (0.7 % + 2)
3
± (2.0 % + 4)
unspecified
1000 V
1 V
unspecified
unspecified
unspecified
unspecified
+1 % + 4
-6 % - 45
Low pass filter
± (0.7 % + 2) ± (1.0 % + 4)
unspecified
1. Below 10 % of range, add 12 counts.
2. The Meter is a true rms responding meter. When the input leads are shorted together in the ac functions, the Meter
may display a residual reading between 1 and 30 counts. A 30 count residual reading will cause only a 2-digit
change for readings over 3 % of range. Using REL to offset this reading may produce a much larger constant error
in later measurements.
3. Frequency range: 1 kHz to 2.5 kHz.
4. A residual reading of up to 13 digits with leads shorted, will not affect stated accuracy above 3 % of range.
5. Specification increases from -1% at 200 Hz to -6% at 440 Hz when filter is in use.
Table 3. Models 83 and 88 AC Voltage Function Specifications
Function
Range
Resolution
Accuracy
50 Hz - 60 Hz
30 Hz - 1 kHz
1 kHz - 5 kHz
600.0 mV
6.000 V
60.00 V
600.0 V
1000 V
0.1 mV
0.001 V
0.01 V
0.1 V
± (0.5 % + 4)
± (0.5 % + 2)
± (0.5 % + 2)
± (0.5 % + 2)
± (0.5 % + 2)
± (1.0 % + 4)
± (1.0 % + 4)
± (1.0 % + 4)
± (1.0 % + 4)
± (1.0 % + 4)
± (2.0 % + 4)
± (2.0 % + 4)
± (2.0 % + 4)
± (2.0 % + 4)
unspecified
K1
2
1 V
1. Below a reading of 200 counts, add 10 counts.
2. Frequency range: 1 kHz to 2.5 kHz.
For models 83 and 88, ac conversions are ac-coupled and are average- responding, rms-indicating.
5
Download from Www.Somanuals.com. All Manuals Search And Download.
80 Series V
Calibration Manual
Table 4. DC Voltage, Resistance, and Conductance Function Specifications
Accuracy
Function
Range
Resolution
Model 83
Model 87
Model 88
6.000 V
0.001 V
0.01 V
0.1 V
1 V
± (0.1 % + 1)
± (0.1 % + 1)
± (0.1 % + 1)
± (0.1 % + 1)
± (0.05 % + 1)
± (0.05 % + 1)
± (0.05 % + 1)
± (0.05 % + 1)
± (0.1 % + 1)
± (0.1 % + 1)
± (0.1 % + 1)
± (0.1 % + 1)
L
60.00 V
600.0 V
1000 V
F
600.0 mV
0.1 mV
± (0.3 % + 1)
± (0.1 % + 1)
± (0.3 % + 1)
mV
1
1
1
600.0 Ω
0.1 Ω
± (0.4 % + 2)
± (0.4 % + 1)
± (0.4 % + 1)
± (0.7 % + 1)
± (0.7 % + 1)
± (1.0 % + 3)
± (0.2 % + 2)
± (0.4 % + 2)
± (0.4 % + 1)
± (0.4 % + 1)
± (0.7 % + 1)
± (0.7 % + 1)
± (1.0 % + 3)
e
6.000 kΩ
60.00 kΩ
600.0 kΩ
6.000 MΩ
50.00 MΩ
60.00 nS
0.001 kΩ
0.01 kΩ
0.1 kΩ
0.001 MΩ
0.01 MΩ
0.01 nS
± (0.2 % + 1)
± (0.2 % + 1)
± (0.6 % + 1)
± (0.6 % + 1)
2
2
2
± (1.0 % + 3)
1
1
1
nS
± (1.0 % + 10)
± (1.0 % + 10)
± (1.0 % + 10)
1. When using the REL ∆ function to compensate for offsets.
2. Add 0.5 % of reading when measuring above 30 MΩ in the 50 MΩ range, and 20 counts below 33 nS in the 60
nS range.
Table 5. Temperature Specifications (87 and 88 Only)
Temperature
-200 °C to +1090 °C
-328 °F to +1994 °F
Resolution
0.1 °C
0.1 °F
Accuracy1,2
1 % + 10
1 % + 18
1. Does not include error of the thermocouple probe.
2. Accuracy specification assumes ambient temperature stable to ± 1 °C. For ambient temperature changes of ± 5 °C,
rated accuracy applies after 1 hour.
Table 6. Current Function Specifications
Accuracy
Burden
Voltage
(typical)
Function
Range
Resolution
1
2, 3
1
Model 83
Model 87
Model 88
5
5
5
5
5
5
5
5
60.00 mA
400.0 mA
6.000 A
0.01 mA
0.1 mA
0.001 A
0.01 A
± (1.2 % + 2)
± (1.2 % + 2)
± (1.2 % + 2)
± (1.2 % + 2)
± (1.0 % + 2)
± (1.0 % + 2)
± (1.0 % + 2)
± (1.0 % + 2)
± (1.2 % + 2)
± (1.2 % + 2)
± (1.2 % + 2)
± (1.2 % + 2)
1.8 mV/mA
1.8 mV/mA
0.03 V/A
mA
6
6
\
(45 Hz to
2 kHz)
4
10.00 A
0.03 V/A
60.00 mA
400.0 mA
6.000 A
0.01 mA
0.1 mA
0.001 A
0.01 A
± (0.4 % + 4)
± (0.4 % + 2)
± (0.4 % + 4)
± (0.4 % + 2)
± (0.2 % + 4)
± (0.2 % + 2)
± (0.2 % + 4)
± (0.2 % + 2)
± (0.4 % + 4)
± (0.4 % + 2)
± (0.4 % + 4)
± (0.4 % + 2)
1.8 mV/mA
1.8 mV/mA
0.03 V/A
mA
[
4
10.00 A
0.03 V/A
µA B
(45 Hz to
2 kHz)
5
5
600.0 µA
6000 µA
0.1 µA
1 µA
± (1.2 % + 2)
± (1.2 % + 2)
± (1.0 % + 2)
± (1.0 % + 2)
± (1.2 % + 2)
± (1.2 % + 2)
100 µV/µA
100 µV/µA
5
5
± (0.2 % + 4)
± (0.2 % + 2)
600.0 µA
6000 µA
0.1 µA
1 µA
± (0.4 % + 4)
± (0.4 % + 2)
± (0.4 % + 4)
± (0.4 % + 2)
100 µV/µA
100 µV/µA
µAF
1. AC conversion for Model 83 and 88 is ac coupled and calibrated to the rms value of a sine wave input.
2. AC conversions for Model 87 are ac coupled, true rms responding, and valid from 3 % to 100 % of range, except
400 mA range (5 % to 100 % of range) and 10 A range (15 % to 100 % or range).
3. Model 87 is a true rms responding meter. When the input leads are shorted together in the ac functions, the Meter
may display a residual reading between 1 and 30 counts. A 30 count residual reading will cause only a 2 digit
change for readings over 3 % of range. Using REL to offset this reading may produce a much larger constant
error in later measurements
4. W10 A continuous up to 35 °C; < 20 minutes on, 5 minutes off at 35 °C to 55 °C. 20 A for 30 seconds maximum;
> 10 A unspecified.
5. Below a reading of 200 counts, add 10 counts.
6. 400 mA continuous; 600 mA for 18 hrs maximum.
6
Download from Www.Somanuals.com. All Manuals Search And Download.
Digital Multimeter
Specifications
Table 7. Capacitance and Diode Function Specifications
Function
Range
Resolution
Accuracy
E
10.00 nF
100.0 nF
1.000 µF
10.00 µF
100.0 µF
9999 µF
0.01 nF
0. 1 nF
0.001 µF
0.01 µF
0.1 µF
1 µF
± (1 % + 2)1
± (1 % + 2)1
± (1 % + 2)
± (1 % + 2)
± (1 % + 2)
± (1 % + 2)
G
3.000 V
0.001 V
± (2 % + 1)
1. With a film capacitor or better, using Relative mode to zero residual.
Table 8. Frequency Counter Specifications (Models 87 and 83)
Function
Frequency
(0.5 Hz to 200 kHz,
pulse width > 2 µs)
Range
199.99
Resolution
Accuracy
0.01 Hz
0.1 Hz
0.001 kHz
0.01 kHz
0.1 kHz
± (0.005 % + 1)
± (0.005 % + 1)
± (0.005 % + 1)
± (0.005 % + 1)
unspecified
1999.9
19.999 kHz
199.99 kHz
> 200 kHz
Table 9. Frequency Counter Specifications (Model 88)
Pulse Width
Resolution
(ms)
Function
Range
199.99
Resolution
0.01 Hz
Accuracy
1
Range (ms)
2
Frequency
± (0.01 % + 1)
± (0.01 % + 1)
± (0.01 % + 1)
± (0.01 % + 1)
unspecified
1999.9
5.00
0.500
0.0500
0.1
0.01
0.001
0.0001
(0.5 Hz to 200 kHz,
pulse width > 2 µs)
1999.9
0.1 Hz
19.999 kHz
199.99 kHz
> 200 kHz
0.001 kHz
0.01 kHz
0.1 kHz
RPM n
RPM o
30 to 9,000
1 RPM
1 RPM
± 2 RPM
± 2 RPM
60 to 12,000
3
% Duty Cycle
0.0 to 99.9% (0.5 Hz to 200 kHz, Pulse Width >2 µs)
3
Pulse Width
0.002 to 1999.9 ms (4 Hz to 200 kHz, Pulse Width >2 µs)
1. Pulse Width range is determined by the frequency of the signal.
2. Frequency measurements can be made on voltage or current inputs. The current inputs are always dc coupled.
3. For rise times >1 µs. Duty Cycle Accuracy: ±(0.2 % per kHz + 0.1%). Pulse Width Accuracy: ±(0.002 ms + 3).
Table 10. Frequency Counter Sensitivity and Trigger Levels
Minimum Sensitivity (RMS Sine wave)
Approximate Trigger Level
(DC Voltage Function)
1
Input Range
5 Hz - 20 kHz
0.5 Hz - 200 kHz
600 mV dc
600 mV ac
6 V
70 mV (to 400 Hz)
70 mV (to 400 Hz)
150 mV
40 mV
150 mV
0.3 V
3 V
0.7 V
1.7 V
4 V
60 V
7 V (≤140 kHz)
70 V (≤14.0 kHz)
200 V (≤1.4 kHz)
600 V
30 V
40 V
100 V
1000 V
100 V
Duty Cycle Range
Accuracy
Within ± (0.2% per kHz + 0.1 %) for rise times < 1 µs.
0.0 to 99.9 %
1. Maximum input for specified accuracy = 10X Range or 1000 V.
7
Download from Www.Somanuals.com. All Manuals Search And Download.
80 Series V
Calibration Manual
Table 11. Electrical Characteristics of the Terminals
Input
Common Mode
Rejection Ratio
(1 kΩ unbalance)
Overload
Impedance
(nominal)
Function
Normal Mode Rejection
> 60 dB at 50 Hz or 60 Hz
> 60 dB at 50 Hz or 60 Hz
1
Protection
L
1000 V rms
1000 V rms
1000 V rms
10 MΩ < 100 pF
10 MΩ < 100 pF
> 120 dB at dc,
50 Hz or 60 Hz
F
> 120 dB at dc,
50 Hz or 60 Hz
mV
K
10 MΩ < 100 pF
> 60 dB, dc to 60 Hz
(ac-coupled)
Full Scale Voltage
Typical Short Circuit Current
Open Circuit
Test Voltage
50 MΩ or
To 6.0 MΩ
600 Ω
6 k
60 k
600 k
6 M 50 M
60 nS
e
1000 V rms
1000 V rms
< 7.9 V dc
< 3.9 V dc
< 4.1 V dc
< 4.5 V dc
1 mA 100 µA 10 µA
1µA
1 µA 0.5 µA
G
3.000 V dc
0.6 mA typical
6
10 V Hz max
Table 12. MIN MAX Recording Specifications
Accuracy
Model
Nominal Response
83
100 ms to 80 %
Specified accuracy ± 12 counts for changes > 200 ms in duration
(± 40 counts in ac with beeper on)
87, 88
100 ms to 80 %
(dc functions)
Specified accuracy ± 12 counts for changes > 200 ms in duration
120 ms to 80 %
(ac functions)
Specified accuracy ± 40 counts for changes > 350 ms and inputs > 25 % of
range
250 µs (peak)1
Specified accuracy ± 100 counts for changes > 250 µs in duration
(add ± 100 counts for readings over 6000 counts)
(add ± 100 counts for readings in Low Pass mode)
1. For repetitive peaks: 1 ms for single events.
8
Download from Www.Somanuals.com. All Manuals Search And Download.
Digital Multimeter
Basic Maintenance
Basic Maintenance
XWWarning
To avoid possible electric shock or personal injury:
•
Remove the test leads and any input signals before opening
the case or replacing the battery or fuses.
•
Repairs or servicing covered in this manual should be
performed only by qualified personnel.
Cleaning the Meter
XWWarning
To avoid possible electric shock, personal injury, or damage to
the meter, never allow water inside the case.
WCaution
To avoid damaging the Meter, never apply abrasives, solvents,
aromatic hydrocarbons, chlorinated solvents, or methanol-
based fluids to the Meter.
Periodically wipe the Meter case with Fluke “MeterCleanerTM” or a damp cloth and mild
detergent.
Dirt or moisture in the A or mA µA input terminals can affect readings and can falsely
activate the Input Alert feature without the test leads being inserted. Such contamination
may be dislodged by turning the Meter over and, with all test leads removed, gently
tapping on the case.
Thoroughly clean the terminals as follows:
1. Turn the Meter off and remove all test leads.
2. Soak a clean swab with isopropyl alcohol and work the swab around in each input
terminal to remove contaminates.
Opening the Meter Case
WCaution
To avoid unintended circuit shorting, always place the
uncovered Meter assembly on a protective surface. When the
case of the Meter is open, circuit connections are exposed.
To open the Meter case, refer to Figure 1 and do the following:
1. Disconnect test leads from any live source, turn the rotary knob to OFF, and remove
the test leads from the front terminals.
2. Remove the battery door by using a flat-blade screwdriver to turn the battery door
screws 1/4-turn counterclockwise.
3. The case bottom is secured to the case top by three screws and two internal snaps (at
the LCD end). Using a Phillips-head screwdriver, remove the three screws.
9
Download from Www.Somanuals.com. All Manuals Search And Download.
80 Series V
Calibration Manual
WCaution
To avoid damaging the Meter, the gasket that is sealed to the
bottom case, and is between the two case halves, must remain
with the case bottom. The case top lifts away from the gasket
easily. Do not damage the gasket or attempt to separate the
case bottom from the gasket.
4. Hold the Meter display side up.
5. Pushing up from the inside of the battery compartment, disengage the case top from
the gasket.
6. Gently unsnap the case top at the display end, see Figure 1.
F1
F2
ama12.eps
Figure 1. Opening the Meter, Battery and Fuse Replacement
Accessing the PCA and Replacing the LCD
Once the case has been opened, the A1 Main PCA can easily be removed. The shields
disconnect from the PCA as follows:
1. Remove the five Phillips-head screw securing the top and bottom shields to the PCA.
2. Remove the top shield assembly that also houses the LCD and lightpipe for the LCD
backlight.
3. To access the LCD, unsnap the LCD mask using a small flat-blade screwdriver. The
LCD may now be removed. Refer to Figure 2.
Note
Two elastomeric connectors make electrical contact between the LCD and
the PCA. These connectors usually stick to the LCD when it is removed. If
the connectors are to be reused, do not handle them, as the electrical
contact points might become contaminated. Use tweezers to remove these
connectors.
4. To reinstall the connectors, replace the LCD and LCD mask and lay the top shield
face down. Install the elastomeric connector strips into the slots on the top shield.
10
Download from Www.Somanuals.com. All Manuals Search And Download.
Digital Multimeter
Basic Maintenance
5. Place the PCA onto the top shield so that the screw holes align.
6. Place the bottom shield onto the PCA and secure the assembly with five Phillips-head
screws. Ensure that the shields are tightly attached. Properly fitted shields are
required for the Meter to perform to specifications.
LCD Mask
LCD
ama08f.eps
Figure 2. Removing LCD Mask to Access LCD
Reassembling the Meter Case
To reassemble the Meter case:
1. Verify that the rotary knob and circuit board switch are in the OFF position, and that
the gasket remains secured to the bottom case.
2. Place the PCA into the bottom case.
3. Place the case top on the case bottom.
4. To avoid damaging the battery wire, ensure the wire exits the middle of the battery
compartment.
5. Properly seat the case gasket and snap the case halves together above the LCD end.
See Figure 1.
6. Reinstall the three case screws and the battery door.
7. Secure the battery door by turning the screw 1/4-turn clockwise.
8. Go to “Performance Tests” later in this document, and perform the procedures
described.
11
Download from Www.Somanuals.com. All Manuals Search And Download.
80 Series V
Calibration Manual
Replacing the Battery
Replace the battery with a 9-V battery (NEDA A1604, 6F22, or 006P).
XWWarning
To avoid false readings, which could lead to possible electric
shock or personal injury, replace the battery as soon as the
battery indicator (b) appears. If the display shows “bAtt” the
Meter will not function until the battery is replaced.
Replace the battery as follows, refer to Figure 1:
1. Turn the rotary knob to OFF and remove the test leads from the terminals.
2. Remove the battery door by using a standard-blade screwdriver to turn the battery
door screws one-quarter turn counterclockwise.
3. Remove the old battery and replace it with a new one.
4. Align the battery leads so that they not pinched between the battery door and the case
bottom.
5. Secure the door by turning the screws one-quarter turn clockwise.
Testing Fuses and Current Circuitry
If a test lead is plugged into the mA/µA or A terminal and the rotary knob is turned to a
non-current function, the Meter chirps and flashes “LEAd” if the fuse associated with that
current terminal is good. If the Meter does not chirp or flash “LEAd”, the fuse is bad and
must be replaced. Refer to Table 17 for the appropriate replacement fuse.
After replacing the fuse, use the following procedure to verify the integrity of the new
fuse and the current circuitry. Refer to Figure 3.
1. Turn the rotary knob to N.
2. To test F2, insert a test lead into the Iinput terminal and touch the probe to
the A input terminal.
Note
The input receptacles contain split contacts. Be sure to touch the probe to
the half of the receptacle nearest the LCD.
3. The display should indicate between 00.0 Ω and 00.5 Ω. If the display reads OL,
replace the fuse and test again. If the display reads another value, further servicing is
required.
4. To test F1, move the probe from the A input terminal to the mA/UA input terminal.
5. The display should read between 0.995 kΩ and 1.005 kΩ. If the display reads OL,
replace the fuse and test again. If the display reads another value, further servicing is
required.
XWWarning
To avoid electrical shock or personal injury:
•
Remove the test leads and any input signals before
replacing the battery or fuses.
•
Install ONLY specified replacement fuses with the
amperage, voltage, and speed ratings shown in Table 17.
12
Download from Www.Somanuals.com. All Manuals Search And Download.
Digital Multimeter
Required Equipment
TRUE RMS MULTIMETER
87
V
Good F2 fuse: 00.0 Ω to
00.5 Ω
C/
F
˚
˚
MIN MAX
RANGE
AutoHOLD
Peak MIN MAX
Replace fuse: OL
REL
Hz
%
4½ DIGITS
1
Second
mV
mA
A
V
LO
A
V
OFF
Touch top half
of input contacts
A
mA
A
COM
V
400mA
FUSED
10A MAX
FUSED
TRUE RMS MULTIMETER
87
V
C/
F
˚
˚
MIN MAX
Peak MIN MAX
RANGE
AutoHOLD
REL
Hz
%
4½ DIGITS
1
Second
mV
mA
A
Good F1 fuse: 0.995 kΩ to
1.005 kΩ
V
LO
A
V
OFF
A
mA
A
COM
V
Replace fuse: OL
400mA
FUSED
10A
FU
aom5f.eps
Figure 3. Testing the Current Input Fuses
Replacing the Fuses
To replace the fuse(s), perform the following procedure.
1. To open the Meter, refer to “Opening the Meter Case”. See Figure 1.
2. Grasp the fuse in the center with needle nose pliers. Pull straight up on the fuse to
remove it from the fuse clips.
3. Install ONLY specified replacement fuses with the amperage, voltage, and speed
ratings shown in Table 17.
4. To close the Meter, refer to “Reassembling the Meter Case”.
Required Equipment
Required equipment for the performance tests is listed in Table 13. If the recommended
models are not available, equipment with equivalent specifications may be used.
XWWarning
•
•
To avoid shock or injury, do not perform the verification
tests or calibration adjustment procedures described in this
manual unless you are qualified to do so.
Repairs or servicing should be performed only by qualified
personnel.
13
Download from Www.Somanuals.com. All Manuals Search And Download.
80 Series V
Calibration Manual
Table 13. Required Equipment
Required Characteristics
Equipment
Calibrator
Recommended Model
AC Voltage Range: 0 - 1000 V ac
Accuracy: ± 0.12 %
Fluke 5520A Multi-Product
Calibrator or equivalent
Frequency Range: 60 - 20000 Hz
Accuracy: ± 3 %
DC Voltage Range: 0 - 1000 V dc
Accuracy: ± 0.012 %
Current Range: 350 µA - 2 A
Accuracy: AC (60 Hz to 1 kHz): ± 0.25 %
DC: ± 0.05 %
Frequency Source: 19.999 kHz - 199.99 kHz
Accuracy: ± 0.0025 %
Amplitude: 150 mV to 6V rms
Accuracy: ± 5 %
Range: 1 Ω - 100 MΩ
Accuracy: 0.065 %
Function Generator
Frequency = 900 kHz
Amplitude = 8.3V
HP33120
Burst mode = 1
Burst rate = 100Hz
Burst Phase = -90 degrees
Fluke 80 AK
K-type
Fluke 80 AK
TC Adapter Accessory
K-type Thermocouple
K-type, mini-plug on both ends
Performance Tests
The following performance tests verify the complete operability of the Meter and check
the accuracy of each Meter function against the Meter’s specifications. Performance tests
should be performed annually to ensure that the Meter is within accuracy specifications.
Accuracy specifications are valid for a period of one year after calibration adjustment,
when measured at an operating temperature of 18 °C to 28 °C and at a maximum of 90 %
relative humidity.
To perform the following tests, it is not necessary to open the case. No adjustments are
necessary. Make the required connections, apply the designated inputs, and determine if
the reading on the Meter display falls within the acceptable range indicated.
Note
If the Meter fails any of these tests, it needs calibration adjustment or repair.
14
Download from Www.Somanuals.com. All Manuals Search And Download.
Digital Multimeter
Performance Tests
Basic Operability Tests
Refer to the following sections to test the basic operability of the Meter.
Testing the Fuses
Refer to “Testing the Fuses”.
Testing the Display
Turn the Meter on while holding down Dto view all segments of the display.
Compare the display with the appropriate examples in Figure 4 and Table 14.
ayi04.eps
Figure 4. Display Features
Table 14. Display Features
Number
Feature
Indication
Y
Polarity indicator for the analog bar graph.
Positive or negative slope indicator for Hz/duty cycle triggering.
The continuity beeper is on.
A
TrigY
X
B
C
D
W
Relative (REL) mode is active.
g
Smoothing is active.
Indicates negative readings. In relative mode, this sign indicates that the
present input is less than the stored reference.
-
E
F
Indicates the presence of a high voltage input. Appears if the input voltage is
30 V or greater (ac or dc). Also appears in low pass filter mode. Also appears
in cal, Hz, and duty cycle modes.
Z
RS
S
G
H
I
J
K
AutoHOLD is active.
Display Hold is active.
p
Indicates the Meter is in Peak Min Max mode and the response time is 250 µs
Indicators for minimum-maximum recording mode.
m MAX
MIN AVG
K
Low pass filter mode.
15
Download from Www.Somanuals.com. All Manuals Search And Download.
80 Series V
Calibration Manual
Table 13. Display Features (cont.)
Indication
Number
Feature
The battery is low. XWWarning: To avoid false readings, which could lead to
possible electric shock or personal injury, replace the battery as soon as the
battery indicator appears.
b
L
M
Amperes (amps), Microamp, Milliamp
Volts, Millivolts
A, µA, mA
V, mV
Microfarad, Nanofarad
µF, nF
nS
Nanosiemens
Percent. Used for duty cycle measurements.
Ohm, Megohm, Kilohm
%
e, Me, ke
Hz, kHz
AC DC
°C, °F
Hertz, Kilohertz
Alternating current, direct current
Degrees Celsius, Degrees Fahrenheit
Displays selected range
N
O
610000 mV
The Meter is in high resolution
(Hi Res) mode. HiRes=19,999
P
HiRes
The Meter is in autorange mode and automatically selects the range with the
best resolution.
Auto
Q
R
The Meter is in manual range mode.
Manual
The number of segments is relative to the full-scale value of the selected
range. In normal operation 0 (zero) is on the left. The polarity indicator at the
left of the graph indicates the polarity of the input. The graph does not operate
with the capacitance, frequency counter functions, temperature, or peak min
max. For more information, see “Bar Graph”. The bar graph also has a zoom
function, as described under "Zoom Mode".
RPMon
oconventional (4 cycle) Counts every other revolution.
nwaste spark of 2 cycle. Counts every revolution.
S
--
Overload condition is detected.
0L
Error Messages
bAtt
Replace the battery immediately.
diSC
In the capacitance function, too much electrical charge is present on the capacitor being tested.
EEPr
Err
Invalid EEPROM data. Have Meter serviced.
Invalid calibration data. Calibrate Meter.
CAL
Err
WTest lead alert. Displayed when the test leads are in the A or mA/µA terminal and the selected
rotary switch position does not correspond to the terminal being used.
LEAd
16
Download from Www.Somanuals.com. All Manuals Search And Download.
Digital Multimeter
Performance Tests
Testing the Pushbuttons
To test the pushbuttons
1. Turn the Meter rotary knob to J.
2. Press each button and note that the meter responds with a beep for each button
press.
3. Press and hold Ba second time to exit MIN MAX mode.
Testing Meter Accuracy
Perform the accuracy test steps in Table 15.
Table 15. Accuracy Tests
Display Reading
83 and 88
Test
Step
Range
5500A Output
Function
87
1
600 mV
600 mV
6 V
330 mV, 60 Hz
327.9 to 332.1
N/A
327.3 to 332.7
586.0 to 614.0
3.275 to 3.325
3.214 to 3.386
32.75 to 33.25
32.14 to 33.86
327.5 to 332.5
323.0 to 337.0
494 to 506
K
AC Volts
2
600 mV, 13 kHz
3.3 V, 60 Hz
3
3.281 to 3.319
N/A
4
6 V
3.3 V, 20 kHz
33 V, 60 Hz
5
60 V
32.81 to 33.19
N/A
6
60 V
33 V, 20 kHz
7
600 V
600 V
1000 V
1000 V
600 mV
600 mV
330 V, 60 Hz
328.1 to 331.9
N/A
8
330 V, 2.5 kHz
500 V, 60 Hz
9
495 to 505
986 to 1014
99.93 to 99.97
199.48 to 199.52
10
11
12
1000 V, 1 kHz
150 mV, 99.95 kHz
150 mV, 199.50 kHz
986 to 1014
99.93 to 99.97
199.48 to 199.52
K
AC Volts
Hz
Frequency
13
14
15
Sensitivity
6 V
0.7 V, 99.95 kHz
7 V, 99.95 kHz
99.93 to 99.97
99.93 to 99.97
999.8 to 1000.2
99.93 to 99.97
99.93 to 99.97
999.8 to 1000.2
60 V
6 V
3.4 V, 1 kHz Sq. Wave
VHz
Trigger level
16
6 V
5 V, 1 kHz, DC offset 2.5 V
Sq. Wave
49.7% to 50.3 %
49.7 % to 50.3 %
VHz
Duty Cycle
17
18
6V
3.3 V dc
33 V dc
3.297 to 3.303
32.97 to 33.03
3.296 to 3.304
32.96 to 33.04
V
DC Volts
60 V
19
20
21
22
600 V
330 V dc
329.7 to 330.3
998 to 1002
329.6 to 330.4
998 to 1002
1000 V
600 mV
600 mV
1000 V dc
33 mV dc
330 mV dc
32.9 to 33.1
32.8 to 33.2
mV
DC Volts
329.6 to 330.4
328.9 to 331.1
17
Download from Www.Somanuals.com. All Manuals Search And Download.
80 Series V
Calibration Manual
Table 14. Accuracy Tests (cont.)
5500A Output
Display Reading
83 and 88
Test
Function
Step
Range
87
23
24
25
26
27
28
29
30
31
e
600 Ω
6 kΩ
330 Ω ( Use 2 wire Comp)1
3.3 kΩ (Use 2 wire Comp)1
329.1 to 330.9
3.292 to 3.308
32.92 to 33.08
327.9 to 332.1
3.279 to 3.321
29.67 to 30.33
- 0.30 to 0.30
9.60 to 10.40
2.939 to 3.061
328.5 to 331.5
3.286 to 3.314
32.86 to 33.14
327.6 to 332.4
3.276 to 3.324
29.67 to 30.33
- 0.30 to 0.30
9.60 to 10.40
2.939 to 3.061
Ohms
60 kΩ
600 kΩ
6 MΩ
50 MΩ
60 nS
60 nS
6 V
33 kΩ
330 kΩ
3.3 MΩ
30 MΩ
nS
Open input
100 MΩ
Conductance
3.0 V dc
G
Diode
32
6 A
3.0 A, 60 Hz
3.0 A
2.962 to 3.038
2.984 to 3.016
9.94 to 10.06
2.968 to 3.032
2.990 to 3.010
9.96 to 10.04
?
AC Amps
33
6 A
A
DC Amps
10A6
10A
33B
A
DC Amps
34
35
36
37
38
39
60 mA
33 mA, 60 Hz
330 mA, 60 Hz
33 mA
32.58 to 33.42
325.8 to 334.2
32.83 to 33.17
328.5 to 331.5
325.8 to 334.2
3258 to 3342
32.65 to 33.35
326.5 to 333.5
32.89 to 33.11
329.1 to 330.9
326.5 to 333.5
3265 to 3335
?
m
400 mA
60 mA
AC Milliamps
m[
DC Milliamp
400 mA
600 µA
6000 µA
330 mA
330 µA, 60 Hz
3300 µA, 60 Hz
?
AC
U
Microamps
40
41
U[
600 µA
330 µA
328.3 to 331.7
3285 to 3315
328.9 to 331.1
3291 to 3309
DC
Microamps
6000 µA
3300 µA
42
43
44
45
46
E
10 nf
Open input2
5 nf5
0.21 to 0.31
04.7 to 05.3
09.2 to 09.8
N/A
0.21 to 0.31
04.7 to 05.3
09.2 to 09.8
376 to 408
226 to 3404
Capacitance
100 nf
100 µf
1000 V
1000 V
9.5 µf
400 V, 400 Hz
400 V, 800 Hz4
K
Low Pass
N/A
Filter
18
Download from Www.Somanuals.com. All Manuals Search And Download.
Digital Multimeter
Performance Tests
Table 14. Accuracy Tests (cont.)
5500A Output
Display Reading
83 and 88
Test
Function
Step
Range
87
6 V dc
8 Vpp, 2 kHz Sq.
Wave, DC offset 2 V
Max = 5.896 to 6.104
Min = -1.898 to -2.102
47
48
L
(87 and 88 only)
Peak Min/Max
0 °C
-1.0 to 1.0
98.0 to 102.0
-1.0 to 1.0
49
50
L
m
(87 and 88 only)
Temperature 3
100 °C
98.0 to 102.0
Backlight
Press backlight button
Press backlight button
Press backlight button
Backlight comes on
51
52
53
Backlight Intensifies
Backlight off
1. Or short test leads and use REL to offset test lead resistance.
2. Remove test leads from unit.
3. To ensure accurate measurement, the Meter and thermocouple adapter must be at the same temperature. After
connecting the thermocouple adapter to the Meter allow for reading to stabalize before recording display reading.
4. The Meter accuracy is not specified at this input signal frequency with Low-pass filter selected. The display reading
shown, check that the Low-pass filter is active and follows an expected roll-off curve.
5. Use REL to compensate for internal Meter and lead capacitance. Test leads must be disconnected from the
calibrator before using REL.
6. W10 A continuous up to 35 °C; < 20 minutes on, 5 minutes off at 35 °C to 55 °C. 20 A for 30 seconds maximum;
> 10 A unspecified.
Testing the Inductive Pickup (88 Only)
To test the inductive pickup, a function generator output will simulate automobile spark
plug signals on a loop of wire containing a 10 eresistor. The pickup will be clamped to
the wire and output voltage from the pickup will be monitored by an oscilloscope.
Perform the following procedure to test the inductive pickup:
1. Solder a 10-inch piece of 14 AWG wire to one end of a 10 e1 % resistor.
2. Connect the other end of the 10 eresistor to the terminal LOW output of the
function generator. Place the other end of the 14-AWG wire to the HIGH output of
the function generator. See Figure 5.
3. Connect a 10X scope probe from channel 2 (dc-coupled) of the oscilloscope across
the 10 eresistor.
4. Clamp the inductive pickup to the wire loop on the HIGH side of the resistor as
shown in Figure 5.
Make sure that the jaws of the inductive pickup are closed completely, and that the
side of the inductive pickup that says "SPARK PLUG SIDE" points toward the
HIGH output of the function generator.
5. Connect a 10X scope probe from channel 1 (dc-coupled) of the oscilloscope across
the output of the inductive pickup.
19
Download from Www.Somanuals.com. All Manuals Search And Download.
80 Series V
Calibration Manual
6. Set up the function generator as follows:
Frequency
Amplitude
Burst Mode
Burst Count
Burst Rate
Burst Phase
900 kHz
8.3 V
yes
1
100 Hz
-90 Degrees
7. Set the oscilloscope for 0.5 V/DIV @ 0.5 µs/DIV.
8. Trigger the waveform on channel 2.
9. Adjust the amplitude of the function generator to produce a 3 VP-P triangle wave.
See Figure 6.
10. Set the oscilloscope for 1.0 V/DIV @ 5.0 ms/DIV.
11. Trigger the waveform on Channel 1.
12. Check that the peak voltage is greater than 5.7 V and decays to less than 1.0 V
between pulses. See Figure 6. Record the peak value for later use.
13. Adjust the function generator output so the peak voltage is 6 V.
14. Set the scope for 1.0 ms/DIV and trigger waveform.
15. Check the amplitude after 3.0 ms from the waveform peak, the voltage amplitude is
2.4 V +0.5/-0.8 (1.6 V to 2.9 V). (See Figure 3-5(C).)
16. Re-adjust the function generator output to obtain the value recorded in step 12.
17. Set the scope for 5.0 ms/DIV.
18. Turn the inductive pickup so that "SPARK PLUG SIDE" points along the wire
connected to the LOW output of the function generator. Check that the waveform is
less than 2 V.
20
Download from Www.Somanuals.com. All Manuals Search And Download.
Digital Multimeter
Calibration Adjustment
Calibration Adjustment
The Meter features closed-case calibration adjustment using known reference sources.
The Meter measures the applied reference source, calculates correction factors and stores
the correction factors in nonvolatile memory.
The following sections present the features and Meter pushbutton functions that can be
used during the Calibration Adjustment Procedure. Perform the Calibration Adjustment
Procedure should the Meter fail any performance test listed in Table 15.
Calibration Adjustment Counter
The Meter contains a calibration adjustment counter. The counter is incremented each
time a Calibration Adjustment Procedure is completed. The value in the counter can be
recorded and used to show that no adjustments have been made during a calibration
cycle.
Use the following steps to view the Meter's calibration counter.
1. While holding down B, turn the rotary knob from OFF to VAC. The Meter
should display “ZCAL”.
2. Press Donce to see the calibration counter. For example "n001".
3. Turn the rotary knob to OFF.
Calibration Adjustment Password
To start the Calibration Adjustment Procedure, the correct 4-button password must be
entered. The password can be changed or reset to the default as described in following
paragraphs. The default password is “1234”.
Changing the Password
Use the following steps to change the Meter's password:
1. While holding down B, turn the rotary knob from OFF to VAC. The Meter
displays “ZCAL”.
2. Press Donce to see the calibration counter.
3. Press Dagain to start the password entry. The Meter displays "????".
4. The Meter buttons represent the digit indicated below when entering or changing the
password:
A= 1 B= 2 C= 3 D= 4
H= 5
E= 6 F= 7 G= 8
Press the 4 buttons to enter the old password. If changing the password for the first
time, enter A(1) B(2) C(3) D(4).
5. Press Cto change the password. The Meter displays "----" if the old password is
correct. If the password is not correct, the Meter emits a double beep, displays "????"
and the password must be entered again. Repeat step 4.
6. Press the 4 buttons of the new password.
7. Press Dto store the new password.
23
Download from Www.Somanuals.com. All Manuals Search And Download.
80 Series V
Calibration Manual
Restoring the Default Password
If the calibration password is forgotten, the default password (1234) can be restored using
the following steps.
1. While holding down B, turn the rotary knob from OFF to VAC. The Meter
displays “ZCAL”.
2. Remove the Meter's top case. Leave the PCA in the bottom case. (See “Opening the
Meter Case”.)
XWWarning
To avoid electrical shock or personal injury, remove the test
leads and any input signal before removing the Meter's top
case.
3. Through an access hole provided in the top shield, short across the keypads on the
PCA. See Figure 7. The Meter should beep. The default password is now restored.
4. Replace the Meter's top case and turn the rotary knob to OFF. (See “Reassembling
the Meter Case).
Access Hole For
PCA Keypad
ama01f.eps
Figure 7. Restoring the Default Password
24
Download from Www.Somanuals.com. All Manuals Search And Download.
Digital Multimeter
Meter Buttons Used in the Calibration Steps
The Meter buttons behave as follows when performing the Calibration Adjustment
Procedure. This may be of help determining why a calibration step is not accepted and for
determining the input value without referring to Table 16.
Press and hold to show the measured value. The measurement value is not
A
calibrated so it may not match the input value. This is normal.
B
G
Press and hold to display the required input amplitude.
Press and hold to display the frequency of the required input.
Press to store the calibration value and advance to the next step. This button
is also used to exit calibration mode after the calibration adjustment
sequence is complete.
D
Calibration Adjustment Procedure
Use the following steps to adjust the Meter's calibration. If the Meter is turned off before
completion of the adjustment procedure, the calibration constants are not changed.
1. While holding down B, turn the rotary knob from OFF to VAC. The Meter
displays “ZCAL”.
2. Press Donce to see the calibration counter.
3. Press Dagain to start the password entry. The Meter displays "????".
4. Press 4 buttons to enter the password.
5. Press Dto go to the first calibration step. The Meter displays "C-01" if the
password is correct. If the password is not correct, the Meter emits a double beep,
displays "????" and the password must be entered again. Repeat step 4.
6. Using Table 16, apply the input value listed for each calibration adjustment step. For
each step, position the rotary switch and apply the input to the terminals as indicated
in the table.
7. After each input value is applied, press Dto accept the value and proceed to the
next step (C-02 and so forth).
Notes
After pressing D, wait until the step number advances before changing
the calibrator source or turning the Meter rotary knob.
If the Meter rotary knob is not in the correct position, or if the measured
value is not within the anticipated range of the input value, the Meter emits
a double beep and will not continue to the next step.
Some adjustment steps take longer to execute than others (10 to 15
seconds). For these steps, the Meter will beep when the step is complete.
Not all steps have this feature.
8. After the final step, the display shows "End" to indicate that the calibration
adjustment is complete. Press Dto go to meter mode.
25
Download from Www.Somanuals.com. All Manuals Search And Download.
80 Series V
Calibration Manual
Notes
Set the calibrator to Standby prior to changing the function switch position
and or after completing adjustment of each function.
If the calibration adjustment procedure is not completed correctly, the
Meter will not operate correctly.
Table 16. Calibration Adjustment Steps
Input Value
Function
(Switch Position)
Adjustment
Step
Input Terminal
Fluke 83-V and
88-V
Fluke 87-V
C-01
C-02
C-03
C-04
C-05
C-06
C-07
C-08
C-09
C-10
C-11
C-12
C-13
C-14
C-15
C-16
C-17
C-18
C-19
C-20
600.0 mV, 60 Hz
600.0 mV, 5 kHz
6.000 V, 60 Hz
6.000 V, 5 kHz
60.00 V, 60 Hz
60.00 V, 5 kHz
600.0 V, 60 Hz
600.0 V, 5 kHz
6.000 V
600.0 mV, 60 Hz
600.0 mV, 20 kHz
6.000 V, 60 Hz
6.000 V, 20 kHz
60.00 V, 60 Hz
60.00 V, 20 kHz
600.0 V, 60 Hz
600.0 V, 10 kHz
6.000 V
K
I
(AC Volts)
L
(DC Volts)
60.00 V
60.00 V
600.0 V
600.0 V
600.0 mV
600.0 mV
L
m
(DC Millivolts)
60.00 mV
60.00 mV
600.0
600.0
e
(Ohms)
6.000 k
6.000 k
60.00 k
60.00 k
600.0 k
600.0 k
6.000 M
6.000 M
0.000
0.000
50.0 M
50.0 M
O
C-21
3.000 V
3.000 V
(Diode Test)
C-22
C-23
6.000 A, 60 Hz
6.000 A dc
6.000 A, 60 Hz
6.000 A dc
A
A
(Amps)
C-24
C-25
C-26
C-27
60.00 mA, 60 Hz
400.0 mA, 60 Hz
60.00 mA dc
60.00 mA, 60 Hz
400.0 mA, 60 Hz
60.00 mA dc
mA
(Milliamps)
mA /UA
400.0 mA dc
400.0 mA dc
600.0 µA ac,
60 Hz
µA
C-28
600.0 µA ac, 60 Hz
(Microamps)
C-29
C-30
C-31
6000 µA, 60 Hz
600.0 µA dc
6000 µA dc
6000 µA, 60 Hz
600.0 µA dc
6000 µA dc
26
Download from Www.Somanuals.com. All Manuals Search And Download.
Digital Multimeter
Service and Parts
Service and Parts
Replacement parts are shown in Table 17 and Figure 8. To order parts and accessories,
refer to “Contacting Fluke”.
Table 17. 80 Series V Final Assembly
Description
Part
Number
Ref Des
Qty
AC72
AC72
BT1
Alligator Clip, Black
Alligator Clip, Red
Battery, 9 V
1670652
1670641
2139179
2064217
2074057
943121
803293
832246
448456
817460
674853
824466
1
1
1
1
1
1
1
3
5
2
1
2
BT2
Cable Assy, 9 V Battery Snap
Lightpipe
CR6
F1W
F2W
H2-4
H5-9
J1-2
J3
Fuse, 0.440 A, 1000 V, FAST
Fuse, 11 A, 1000 V, FAST
Screw, Case
Screw, Bottom Shield
Elastomeric Connector
Top Shield Contact
Foot, Non-Skid
MP10,
MP11
MP2
MP4
MP5
MP5
MP5
MP6
MP8
MP9
MP13
MP14
MP15
MP22
Shield, Top
2073906
2074025
2074002
2073992
2115202
2073871
2100482
822643
1
1
Shield, Bottom
Case Top (PAD XFER) With Window (83-5)
Case Top (PAD XFER) with Window (87-5)
Case Top (PAD XFER) with Window (88-5)
Case Bottom
1
1
1
1
1
1
1
1
1
4
Knob, Switch (PAD XFER)
Detent, Knob
Shock Absorber
828541
O-Ring, Input Receptacle
Holster w/ Tilt Stand
831933
2074033
2073938
1567683
Battery Door
MP27-
MP30
Contact RSOB
MP31
MP31
MP31
MP41
Mask, LCD (PAD XFER) (83-5)
Mask, LCD (PAD XFER) (87-5)
Mask, LCD (PAD XFER) (88-5)
Housing, RSOB
2073961
2073950
2112410
2073945
948609
1
1
1
1
2
MP390-
391
Access Door Fastener
NA
Tiltstand
2074040
2105884
855742
1
1
1
1
1
1
1
1
1
S2
Keypad
TL75
TM1
TM2
TM 2
TM3
TM3
Test Lead Set
80 Series V Getting Started Manual (Multi-language)
80 Series V Quick Reference Card
88 V Quick Reference Card
2101973
2101986
2279006
2101999
2278999
2102915
CD ROM (Contains 80 Series V Users Manual )
CD ROM (Contains 88 V Users Manual )
80 Series V Calibration Manual (this manual)
TM4
(not
shown)
U5
LCD, 4.5 DIGIT,TN, Transflective, Bar Graph, OSPR80
2065213
1273113
1
1
MP81
80BK Thermocouple Assembly, K-Type, Beaded, Molded Dual Banana Plug,
Coiled
WTo ensure safety, use exact replacement only.
27
Download from Www.Somanuals.com. All Manuals Search And Download.
80 Series V
Calibration Manual
TL75
Test Lead Set
MP66
MP8
MP31
U5
MP5
J1-2
CR6
MP2
S2
MP9
Alligator Clips
AC72
MP13
MP41
J3
MP81
F2
MP27-30
F1
W
MP4
Holster
MP15
MP14
BT2
H5-9 (5)
MP6
H2-4 (3)
Tilt Stand
MP10-11
TM1
BT1
TM3
MP22
TM2
MP390-391
aom015c.eps
Figure 8. 80 Series V Final Assembly
28
Download from Www.Somanuals.com. All Manuals Search And Download.
|