Fluke Universal Remote 80 Series V User Manual

®
80 Series V  
Digital Multimeter  
Calibration Manual  
PN 2102915  
September 2004, Rev.1, 2/05  
© 2004-2005 Fluke Corporation, All rights reserved. Printed in USA  
All product names are trademarks of their respective companies.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table of Contents  
Title  
Page  
Introduction........................................................................................................ 1  
Contacting Fluke................................................................................................ 1  
Precautions and Safety Information................................................................... 2  
Electrical Symbols ............................................................................................. 4  
Specifications..................................................................................................... 4  
General Specifications................................................................................... 4  
Detailed Specifications.................................................................................. 5  
Basic Maintenance............................................................................................. 9  
Cleaning the Meter ........................................................................................ 9  
Opening the Meter Case ................................................................................ 9  
Accessing the PCA and Replacing the LCD ................................................. 10  
Reassembling the Meter Case........................................................................ 11  
Replacing the Battery .................................................................................... 12  
Testing Fuses and Current Circuitry.............................................................. 12  
Replacing the Fuses....................................................................................... 13  
Required Equipment .......................................................................................... 13  
Performance Tests.............................................................................................. 14  
Basic Operability Tests.................................................................................. 15  
Testing the Fuses........................................................................................... 15  
Testing the Display........................................................................................ 15  
Testing the Pushbuttons................................................................................. 17  
Testing Meter Accuracy ................................................................................ 17  
Testing the Inductive Pickup (88 Only) ........................................................ 19  
Calibration Adjustment...................................................................................... 23  
Calibration Adjustment Counter.................................................................... 23  
Calibration Adjustment Password ................................................................. 23  
Changing the Password............................................................................. 23  
Restoring the Default Password................................................................ 24  
Meter Buttons Used in the Calibration Steps ................................................ 25  
Calibration Adjustment Procedure ................................................................ 25  
Service and Parts................................................................................................ 27  
i
Download from Www.Somanuals.com. All Manuals Search And Download.  
80 Series V  
Calibration Manual  
ii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
List of Tables  
Table  
Title  
Page  
1.  
2.  
3.  
4.  
5.  
6.  
7.  
8.  
9.  
Electrical Symbols ................................................................................................. 4  
Model 87 AC Voltage Function Specifications ..................................................... 5  
Models 83 and 88 AC Voltage Function Specifications........................................ 5  
DC Voltage, Resistance, and Conductance Function Specifications ..................... 6  
Temperature Specifications (87 and 88 Only) ....................................................... 6  
Current Function Specifications............................................................................. 6  
Capacitance and Diode Function Specifications.................................................... 7  
Frequency Counter Specifications (Models 87 and 83)......................................... 7  
Frequency Counter Specifications (Model 88) ...................................................... 7  
10. Frequency Counter Sensitivity and Trigger Levels................................................ 7  
11. Electrical Characteristics of the Terminals............................................................. 8  
12. MIN MAX Recording Specifications .................................................................... 8  
13. Required Tools and Equipment.............................................................................. 14  
14. Display Features..................................................................................................... 15  
15. Accuracy Tests....................................................................................................... 17  
16. Calibration Adjustment Steps................................................................................. 26  
17. 80 Series V Final Assembly................................................................................... 27  
iii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
80 Series V  
Calibration Manual  
iv  
Download from Www.Somanuals.com. All Manuals Search And Download.  
List of Figures  
Figure  
Title  
Page  
1.  
2.  
3.  
4.  
5.  
6.  
7.  
8.  
Opening the Meter, Battery and Fuse Replacement............................................... 10  
Removing LCD Mask to Access LCD................................................................... 11  
Testing the Current Input Fuses ............................................................................. 13  
Display Features..................................................................................................... 15  
Setup for Inductive Pickup Test............................................................................. 21  
Waveform for Inductive Pickup Test..................................................................... 22  
Restoring the Default Password ............................................................................ 24  
80 Series V Final Assembly................................................................................... 28  
v
Download from Www.Somanuals.com. All Manuals Search And Download.  
80 Series V  
Calibration Manual  
vi  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Introduction  
XWWarning  
To avoid shock or injury:  
Read “Precautions and Safety Information” before  
performing the verification tests or calibration adjustment  
procedures documented in this manual.  
Do not perform the verification tests or calibration  
adjustment procedures described in this manual unless you  
are qualified to do so.  
The information provided in this manual is for the use of  
qualified personnel only.  
WCaution  
The 80 Series V Digital Multimeters contains parts that can  
be damaged by static discharge.  
Follow the standard practices for handling static sensitive  
devices.  
The 80 Series V Calibration Manual provides the following information:  
Safety information  
Specifications  
Basic maintenance (cleaning, replacing the battery and fuses)  
Performance test procedures  
Calibration adjustment procedures  
Replaceable parts  
For complete operating instructions, refer to the 80 Series V or 88 Series V Users  
Manual.  
Contacting Fluke  
To contact Fluke or locate the nearest Service Center, call one of the following telephone  
numbers:  
USA: 1-888-44-FLUKE (1-888-443-5853)  
Canada: 1-800-36-FLUKE (1-800-363-5853)  
Europe: +31 402-675-200  
Japan: +81-3-3434-0181  
Singapore: +65-738-5655  
Anywhere in the world: +1-425-446-5500  
1
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
80 Series V  
Calibration Manual  
XWWarning  
To avoid possible electric shock or personal injury, inspect the  
test leads for damaged insulation or exposed metal. Check the  
test leads for continuity. Replace damaged test leads before  
using the Meter.  
Precautions and Safety Information  
In this manual, a Warning identifies conditions and actions that pose hazard(s) to the  
user; a Caution identifies conditions and actions that may damage the Meter or the test  
instruments.  
XWWarning  
To avoid possible electric shock or personal injury, follow these  
guidelines:  
Use this Meter only as specified in this manual or the  
protection provided by the Meter might be impaired.  
Do not use the Meter if it is damaged. Before using the  
Meter, inspect the case. Look for cracks or missing plastic.  
Pay particular attention to the insulation surrounding the  
connectors.  
Make sure the battery door is closed and latched before  
operating the Meter.  
Replace the battery as soon as the battery indicator (M)  
appears.  
Remove test leads from the Meter before opening the  
battery door.  
Inspect the test leads for damaged insulation or exposed  
metal. Check the test leads for continuity. Replace damaged  
test leads before using the Meter.  
Do not apply more than the rated voltage, as marked on the  
Meter, between the terminals or between any terminal and  
earth ground.  
Never operate the Meter with the cover removed or the case  
open.  
Use caution when working with voltages above 30 V ac rms,  
42 V ac peak, or 60 V dc. These voltages pose a shock  
hazard.  
Use only the replacement fuses specified in this manual.  
Use the proper terminals, function, and range for  
measurements.  
Avoid working alone.  
When measuring current, turn off circuit power before  
connecting the Meter in the circuit. Remember to place the  
Meter in series with the circuit.  
2
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Digital Multimeter  
Precautions and Safety Information  
When making electrical connections, connect the common  
test lead before connecting the live test lead; when  
disconnecting, disconnect the live test lead before  
disconnecting the common test lead.  
Do not use the Meter if it operates abnormally. Protection  
may be impaired. When in doubt, have the Meter serviced.  
Do not operate the Meter around explosive gas, vapor, or  
dust.  
Use only a single 9 V battery, properly installed in the Meter  
case, to power the Meter.  
When servicing the Meter, use only specified replacement  
parts.  
When using probes, keep fingers behind the finger guards  
on the probes.  
Do not use the Low Pass Filter option to verify the presence  
of hazardous voltages. Voltages greater than what is  
indicated may be present. Make a voltage measurement  
without the filter to detect the possible presence of  
hazardous voltage, then select the filter function.  
XCaution  
To avoid possible damage to the Meter or to the equipment  
under test, follow these guidelines:  
Disconnect circuit power and discharge all high-voltage  
capacitors before testing resistance, continuity, diodes, or  
capacitance.  
Before measuring current, check the Meter's fuses.  
See " Testing the Fuses”.  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
80 Series V  
Calibration Manual  
Electrical Symbols  
Electrical symbols used on the Meter and in this manual are explained in Table 1.  
Table 1. Electrical Symbols  
AC (Alternating Current)  
DC (Direct Current)  
Hazardous voltage.  
Earth ground  
B
J
Fuse  
F
I
P
Conforms to European Union directives  
X
Risk of Danger. Important  
information. See Manual.  
Conforms to relevant Canadian  
Standards Association directives  
W
M
R
$
T
Battery  
Double insulated  
Capacitance  
Continuity test or continuity beeper  
tone.  
E
G
Underwriters Laboratories  
IEC overvoltage category III  
Diode  
t
IEC overvoltage category IV  
CAT III equipment is designed to  
protect against transients in  
equipment in fixed-equipment  
installations, such as distribution  
panels, feeders and short branch  
circuits, and lighting systems in  
large buildings.  
CAT IV equipment is designed to protect  
against transients from the primary  
supply level, such as an electricity meter  
or an overhead or underground utility  
service.  
CAT III  
CAT IV  
Inspected and licensed by TÜV Product Services.  
s
Specifications  
General Specifications  
Maximum Voltage between any Terminal and Earth Ground: 1000 V rms  
WFuse Protection for mA or µA inputs: 44/100 A, 1000 V FAST Fuse  
WFuse Protection for A input: 11 A, 1000 V FAST Fuse  
Display: Digital: 6000 counts updates 4/sec; (Model 87 and 88 also has 19,999 counts in high-resolution mode).  
Analog Bargraph: 33 segments, updates 40/sec. Frequency: 19,999 counts, updates 3/sec at > 10 Hz  
Temperature: Operating: -20 °C to +55 °C; Storage: -40 °C to +60 °C  
Altitude: Operating: 2000 m; Storage: 10,000 m  
Temperature Coefficient: 0.05 x (specified accuracy)/ °C (< 18 °C or > 28 °C)  
Electromagnetic Compatibility: In an RF field of 3 V/m total accuracy = specified accuracy + 20 counts  
Except: 600 µA dc range total accuracy=specified accuracy + 60 counts.  
Temperature not specified.  
Relative Humidity: 0 % to 90 % (0 °C to 35 °C); 0 % to 7 0% (35 °C to 55 °C)  
Battery Type: 9 V zinc, NEDA 1604 or 6F22 or 006P  
Battery Life: 400 hrs typical with alkaline battery (with backlight off)  
Vibration: Per MIL-PRF-28800 for a Class 2 instrument  
Shock: 1 Meter drop per IEC 61010-1:2001  
Size (HxWxL): 1.25 in x 3.41 in x 7.35 in (3.1 cm x 8.6 cm x 18.6 cm)  
Size with Holster and Flex-Stand: 2.06 in x 3.86 in x 7.93 in (5.2 cm x 9.8 cm x 20.1 cm)  
Weight: 12.5 oz (355 g)  
Weight with Holster and Flex-Stand: 22.0 oz (624 g)  
Safety: Complies with ANSI/ISA S82.01-2004, CSA 22.2 No. 1010.1:2004 to 1000 V Overvoltage Category III, IEC  
4
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Digital Multimeter  
Specifications  
664 to 600 V Overvoltage Category IV. UL listed to UL61010-1. Licensed by TÜV to EN61010-1.  
Detailed Specifications  
For all detailed specifications:  
Accuracy is given as ±([% of reading] + [number of least significant digits]) at 18° C to  
28° C, with relative humidity up to 90 %, for a period of one year after calibration.  
For Model 87 in the 4 ½-digit mode, multiply the number of least significant digits  
(counts) by 10. AC conversions are ac-coupled and valid from 3 % to 100 % of range.  
Model 87 is true rms responding. AC crest factor can be up to 3 at full scale, 6 at half  
scale. For non-sinusoidal wave forms add -(2 % Rdg + 2 % full scale) typical, for a crest  
factor up to 3.  
Table 2. Model 87 AC Voltage Function Specifications  
Function  
Range  
Resolution  
Accuracy  
440 Hz -  
1 kHz  
1
45 - 65 Hz 30 - 200 Hz 200 - 440 Hz  
1 - 5 kHz  
5 - 20 kHz  
600.0 mV  
6.000 V  
60.00 V  
600.0 V  
0.1 mV  
0.001 V  
0.01 V  
0.1 V  
± (0.7 % + 4)  
± (2.0 % + 4)  
± (2.0 % + 20)  
K2,4  
± (1.0 % + 4)  
± (0.7 % + 2)  
3
± (2.0 % + 4)  
unspecified  
1000 V  
1 V  
unspecified  
unspecified  
unspecified  
unspecified  
+1 % + 4  
-6 % - 45  
Low pass filter  
± (0.7 % + 2) ± (1.0 % + 4)  
unspecified  
1. Below 10 % of range, add 12 counts.  
2. The Meter is a true rms responding meter. When the input leads are shorted together in the ac functions, the Meter  
may display a residual reading between 1 and 30 counts. A 30 count residual reading will cause only a 2-digit  
change for readings over 3 % of range. Using REL to offset this reading may produce a much larger constant error  
in later measurements.  
3. Frequency range: 1 kHz to 2.5 kHz.  
4. A residual reading of up to 13 digits with leads shorted, will not affect stated accuracy above 3 % of range.  
5. Specification increases from -1% at 200 Hz to -6% at 440 Hz when filter is in use.  
Table 3. Models 83 and 88 AC Voltage Function Specifications  
Function  
Range  
Resolution  
Accuracy  
50 Hz - 60 Hz  
30 Hz - 1 kHz  
1 kHz - 5 kHz  
600.0 mV  
6.000 V  
60.00 V  
600.0 V  
1000 V  
0.1 mV  
0.001 V  
0.01 V  
0.1 V  
± (0.5 % + 4)  
± (0.5 % + 2)  
± (0.5 % + 2)  
± (0.5 % + 2)  
± (0.5 % + 2)  
± (1.0 % + 4)  
± (1.0 % + 4)  
± (1.0 % + 4)  
± (1.0 % + 4)  
± (1.0 % + 4)  
± (2.0 % + 4)  
± (2.0 % + 4)  
± (2.0 % + 4)  
± (2.0 % + 4)  
unspecified  
K1  
2
1 V  
1. Below a reading of 200 counts, add 10 counts.  
2. Frequency range: 1 kHz to 2.5 kHz.  
For models 83 and 88, ac conversions are ac-coupled and are average- responding, rms-indicating.  
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
80 Series V  
Calibration Manual  
Table 4. DC Voltage, Resistance, and Conductance Function Specifications  
Accuracy  
Function  
Range  
Resolution  
Model 83  
Model 87  
Model 88  
6.000 V  
0.001 V  
0.01 V  
0.1 V  
1 V  
± (0.1 % + 1)  
± (0.1 % + 1)  
± (0.1 % + 1)  
± (0.1 % + 1)  
± (0.05 % + 1)  
± (0.05 % + 1)  
± (0.05 % + 1)  
± (0.05 % + 1)  
± (0.1 % + 1)  
± (0.1 % + 1)  
± (0.1 % + 1)  
± (0.1 % + 1)  
L
60.00 V  
600.0 V  
1000 V  
F
600.0 mV  
0.1 mV  
± (0.3 % + 1)  
± (0.1 % + 1)  
± (0.3 % + 1)  
mV  
1
1
1
600.0 Ω  
0.1 Ω  
± (0.4 % + 2)  
± (0.4 % + 1)  
± (0.4 % + 1)  
± (0.7 % + 1)  
± (0.7 % + 1)  
± (1.0 % + 3)  
± (0.2 % + 2)  
± (0.4 % + 2)  
± (0.4 % + 1)  
± (0.4 % + 1)  
± (0.7 % + 1)  
± (0.7 % + 1)  
± (1.0 % + 3)  
e
6.000 kΩ  
60.00 kΩ  
600.0 kΩ  
6.000 MΩ  
50.00 MΩ  
60.00 nS  
0.001 kΩ  
0.01 kΩ  
0.1 kΩ  
0.001 MΩ  
0.01 MΩ  
0.01 nS  
± (0.2 % + 1)  
± (0.2 % + 1)  
± (0.6 % + 1)  
± (0.6 % + 1)  
2
2
2
± (1.0 % + 3)  
1
1
1
nS  
± (1.0 % + 10)  
± (1.0 % + 10)  
± (1.0 % + 10)  
1. When using the REL function to compensate for offsets.  
2. Add 0.5 % of reading when measuring above 30 Min the 50 Mrange, and 20 counts below 33 nS in the 60  
nS range.  
Table 5. Temperature Specifications (87 and 88 Only)  
Temperature  
-200 °C to +1090 °C  
-328 °F to +1994 °F  
Resolution  
0.1 °C  
0.1 °F  
Accuracy1,2  
1 % + 10  
1 % + 18  
1. Does not include error of the thermocouple probe.  
2. Accuracy specification assumes ambient temperature stable to ± 1 °C. For ambient temperature changes of ± 5 °C,  
rated accuracy applies after 1 hour.  
Table 6. Current Function Specifications  
Accuracy  
Burden  
Voltage  
(typical)  
Function  
Range  
Resolution  
1
2, 3  
1
Model 83  
Model 87  
Model 88  
5
5
5
5
5
5
5
5
60.00 mA  
400.0 mA  
6.000 A  
0.01 mA  
0.1 mA  
0.001 A  
0.01 A  
± (1.2 % + 2)  
± (1.2 % + 2)  
± (1.2 % + 2)  
± (1.2 % + 2)  
± (1.0 % + 2)  
± (1.0 % + 2)  
± (1.0 % + 2)  
± (1.0 % + 2)  
± (1.2 % + 2)  
± (1.2 % + 2)  
± (1.2 % + 2)  
± (1.2 % + 2)  
1.8 mV/mA  
1.8 mV/mA  
0.03 V/A  
mA  
6
6
\
(45 Hz to  
2 kHz)  
4
10.00 A  
0.03 V/A  
60.00 mA  
400.0 mA  
6.000 A  
0.01 mA  
0.1 mA  
0.001 A  
0.01 A  
± (0.4 % + 4)  
± (0.4 % + 2)  
± (0.4 % + 4)  
± (0.4 % + 2)  
± (0.2 % + 4)  
± (0.2 % + 2)  
± (0.2 % + 4)  
± (0.2 % + 2)  
± (0.4 % + 4)  
± (0.4 % + 2)  
± (0.4 % + 4)  
± (0.4 % + 2)  
1.8 mV/mA  
1.8 mV/mA  
0.03 V/A  
mA  
[
4
10.00 A  
0.03 V/A  
µA B  
(45 Hz to  
2 kHz)  
5
5
600.0 µA  
6000 µA  
0.1 µA  
1 µA  
± (1.2 % + 2)  
± (1.2 % + 2)  
± (1.0 % + 2)  
± (1.0 % + 2)  
± (1.2 % + 2)  
± (1.2 % + 2)  
100 µV/µA  
100 µV/µA  
5
5
± (0.2 % + 4)  
± (0.2 % + 2)  
600.0 µA  
6000 µA  
0.1 µA  
1 µA  
± (0.4 % + 4)  
± (0.4 % + 2)  
± (0.4 % + 4)  
± (0.4 % + 2)  
100 µV/µA  
100 µV/µA  
µAF  
1. AC conversion for Model 83 and 88 is ac coupled and calibrated to the rms value of a sine wave input.  
2. AC conversions for Model 87 are ac coupled, true rms responding, and valid from 3 % to 100 % of range, except  
400 mA range (5 % to 100 % of range) and 10 A range (15 % to 100 % or range).  
3. Model 87 is a true rms responding meter. When the input leads are shorted together in the ac functions, the Meter  
may display a residual reading between 1 and 30 counts. A 30 count residual reading will cause only a 2 digit  
change for readings over 3 % of range. Using REL to offset this reading may produce a much larger constant  
error in later measurements  
4. W10 A continuous up to 35 °C; < 20 minutes on, 5 minutes off at 35 °C to 55 °C. 20 A for 30 seconds maximum;  
> 10 A unspecified.  
5. Below a reading of 200 counts, add 10 counts.  
6. 400 mA continuous; 600 mA for 18 hrs maximum.  
6
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Digital Multimeter  
Specifications  
Table 7. Capacitance and Diode Function Specifications  
Function  
Range  
Resolution  
Accuracy  
E
10.00 nF  
100.0 nF  
1.000 µF  
10.00 µF  
100.0 µF  
9999 µF  
0.01 nF  
0. 1 nF  
0.001 µF  
0.01 µF  
0.1 µF  
1 µF  
± (1 % + 2)1  
± (1 % + 2)1  
± (1 % + 2)  
± (1 % + 2)  
± (1 % + 2)  
± (1 % + 2)  
G
3.000 V  
0.001 V  
± (2 % + 1)  
1. With a film capacitor or better, using Relative mode to zero residual.  
Table 8. Frequency Counter Specifications (Models 87 and 83)  
Function  
Frequency  
(0.5 Hz to 200 kHz,  
pulse width > 2 µs)  
Range  
199.99  
Resolution  
Accuracy  
0.01 Hz  
0.1 Hz  
0.001 kHz  
0.01 kHz  
0.1 kHz  
± (0.005 % + 1)  
± (0.005 % + 1)  
± (0.005 % + 1)  
± (0.005 % + 1)  
unspecified  
1999.9  
19.999 kHz  
199.99 kHz  
> 200 kHz  
Table 9. Frequency Counter Specifications (Model 88)  
Pulse Width  
Resolution  
(ms)  
Function  
Range  
199.99  
Resolution  
0.01 Hz  
Accuracy  
1
Range (ms)  
2
Frequency  
± (0.01 % + 1)  
± (0.01 % + 1)  
± (0.01 % + 1)  
± (0.01 % + 1)  
unspecified  
1999.9  
5.00  
0.500  
0.0500  
0.1  
0.01  
0.001  
0.0001  
(0.5 Hz to 200 kHz,  
pulse width > 2 µs)  
1999.9  
0.1 Hz  
19.999 kHz  
199.99 kHz  
> 200 kHz  
0.001 kHz  
0.01 kHz  
0.1 kHz  
RPM n  
RPM o  
30 to 9,000  
1 RPM  
1 RPM  
± 2 RPM  
± 2 RPM  
60 to 12,000  
3
% Duty Cycle  
0.0 to 99.9% (0.5 Hz to 200 kHz, Pulse Width >2 µs)  
3
Pulse Width  
0.002 to 1999.9 ms (4 Hz to 200 kHz, Pulse Width >2 µs)  
1. Pulse Width range is determined by the frequency of the signal.  
2. Frequency measurements can be made on voltage or current inputs. The current inputs are always dc coupled.  
3. For rise times >1 µs. Duty Cycle Accuracy: ±(0.2 % per kHz + 0.1%). Pulse Width Accuracy: ±(0.002 ms + 3).  
Table 10. Frequency Counter Sensitivity and Trigger Levels  
Minimum Sensitivity (RMS Sine wave)  
Approximate Trigger Level  
(DC Voltage Function)  
1
Input Range  
5 Hz - 20 kHz  
0.5 Hz - 200 kHz  
600 mV dc  
600 mV ac  
6 V  
70 mV (to 400 Hz)  
70 mV (to 400 Hz)  
150 mV  
40 mV  
150 mV  
0.3 V  
3 V  
0.7 V  
1.7 V  
4 V  
60 V  
7 V (140 kHz)  
70 V (14.0 kHz)  
200 V (1.4 kHz)  
600 V  
30 V  
40 V  
100 V  
1000 V  
100 V  
Duty Cycle Range  
Accuracy  
Within ± (0.2% per kHz + 0.1 %) for rise times < 1 µs.  
0.0 to 99.9 %  
1. Maximum input for specified accuracy = 10X Range or 1000 V.  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
80 Series V  
Calibration Manual  
Table 11. Electrical Characteristics of the Terminals  
Input  
Common Mode  
Rejection Ratio  
(1 kunbalance)  
Overload  
Impedance  
(nominal)  
Function  
Normal Mode Rejection  
> 60 dB at 50 Hz or 60 Hz  
> 60 dB at 50 Hz or 60 Hz  
1
Protection  
L
1000 V rms  
1000 V rms  
1000 V rms  
10 M< 100 pF  
10 M< 100 pF  
> 120 dB at dc,  
50 Hz or 60 Hz  
F
> 120 dB at dc,  
50 Hz or 60 Hz  
mV  
K
10 M< 100 pF  
> 60 dB, dc to 60 Hz  
(ac-coupled)  
Full Scale Voltage  
Typical Short Circuit Current  
Open Circuit  
Test Voltage  
50 Mor  
To 6.0 MΩ  
600  
6 k  
60 k  
600 k  
6 M 50 M  
60 nS  
e
1000 V rms  
1000 V rms  
< 7.9 V dc  
< 3.9 V dc  
< 4.1 V dc  
< 4.5 V dc  
1 mA 100 µA 10 µA  
1µA  
1 µA 0.5 µA  
G
3.000 V dc  
0.6 mA typical  
6
10 V Hz max  
Table 12. MIN MAX Recording Specifications  
Accuracy  
Model  
Nominal Response  
83  
100 ms to 80 %  
Specified accuracy ± 12 counts for changes > 200 ms in duration  
(± 40 counts in ac with beeper on)  
87, 88  
100 ms to 80 %  
(dc functions)  
Specified accuracy ± 12 counts for changes > 200 ms in duration  
120 ms to 80 %  
(ac functions)  
Specified accuracy ± 40 counts for changes > 350 ms and inputs > 25 % of  
range  
250 µs (peak)1  
Specified accuracy ± 100 counts for changes > 250 µs in duration  
(add ± 100 counts for readings over 6000 counts)  
(add ± 100 counts for readings in Low Pass mode)  
1. For repetitive peaks: 1 ms for single events.  
8
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Digital Multimeter  
Basic Maintenance  
Basic Maintenance  
XWWarning  
To avoid possible electric shock or personal injury:  
Remove the test leads and any input signals before opening  
the case or replacing the battery or fuses.  
Repairs or servicing covered in this manual should be  
performed only by qualified personnel.  
Cleaning the Meter  
XWWarning  
To avoid possible electric shock, personal injury, or damage to  
the meter, never allow water inside the case.  
WCaution  
To avoid damaging the Meter, never apply abrasives, solvents,  
aromatic hydrocarbons, chlorinated solvents, or methanol-  
based fluids to the Meter.  
Periodically wipe the Meter case with Fluke “MeterCleanerTM” or a damp cloth and mild  
detergent.  
Dirt or moisture in the A or mA µA input terminals can affect readings and can falsely  
activate the Input Alert feature without the test leads being inserted. Such contamination  
may be dislodged by turning the Meter over and, with all test leads removed, gently  
tapping on the case.  
Thoroughly clean the terminals as follows:  
1. Turn the Meter off and remove all test leads.  
2. Soak a clean swab with isopropyl alcohol and work the swab around in each input  
terminal to remove contaminates.  
Opening the Meter Case  
WCaution  
To avoid unintended circuit shorting, always place the  
uncovered Meter assembly on a protective surface. When the  
case of the Meter is open, circuit connections are exposed.  
To open the Meter case, refer to Figure 1 and do the following:  
1. Disconnect test leads from any live source, turn the rotary knob to OFF, and remove  
the test leads from the front terminals.  
2. Remove the battery door by using a flat-blade screwdriver to turn the battery door  
screws 1/4-turn counterclockwise.  
3. The case bottom is secured to the case top by three screws and two internal snaps (at  
the LCD end). Using a Phillips-head screwdriver, remove the three screws.  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
80 Series V  
Calibration Manual  
WCaution  
To avoid damaging the Meter, the gasket that is sealed to the  
bottom case, and is between the two case halves, must remain  
with the case bottom. The case top lifts away from the gasket  
easily. Do not damage the gasket or attempt to separate the  
case bottom from the gasket.  
4. Hold the Meter display side up.  
5. Pushing up from the inside of the battery compartment, disengage the case top from  
the gasket.  
6. Gently unsnap the case top at the display end, see Figure 1.  
F1  
F2  
ama12.eps  
Figure 1. Opening the Meter, Battery and Fuse Replacement  
Accessing the PCA and Replacing the LCD  
Once the case has been opened, the A1 Main PCA can easily be removed. The shields  
disconnect from the PCA as follows:  
1. Remove the five Phillips-head screw securing the top and bottom shields to the PCA.  
2. Remove the top shield assembly that also houses the LCD and lightpipe for the LCD  
backlight.  
3. To access the LCD, unsnap the LCD mask using a small flat-blade screwdriver. The  
LCD may now be removed. Refer to Figure 2.  
Note  
Two elastomeric connectors make electrical contact between the LCD and  
the PCA. These connectors usually stick to the LCD when it is removed. If  
the connectors are to be reused, do not handle them, as the electrical  
contact points might become contaminated. Use tweezers to remove these  
connectors.  
4. To reinstall the connectors, replace the LCD and LCD mask and lay the top shield  
face down. Install the elastomeric connector strips into the slots on the top shield.  
10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Digital Multimeter  
Basic Maintenance  
5. Place the PCA onto the top shield so that the screw holes align.  
6. Place the bottom shield onto the PCA and secure the assembly with five Phillips-head  
screws. Ensure that the shields are tightly attached. Properly fitted shields are  
required for the Meter to perform to specifications.  
LCD Mask  
LCD  
ama08f.eps  
Figure 2. Removing LCD Mask to Access LCD  
Reassembling the Meter Case  
To reassemble the Meter case:  
1. Verify that the rotary knob and circuit board switch are in the OFF position, and that  
the gasket remains secured to the bottom case.  
2. Place the PCA into the bottom case.  
3. Place the case top on the case bottom.  
4. To avoid damaging the battery wire, ensure the wire exits the middle of the battery  
compartment.  
5. Properly seat the case gasket and snap the case halves together above the LCD end.  
See Figure 1.  
6. Reinstall the three case screws and the battery door.  
7. Secure the battery door by turning the screw 1/4-turn clockwise.  
8. Go to “Performance Tests” later in this document, and perform the procedures  
described.  
11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
80 Series V  
Calibration Manual  
Replacing the Battery  
Replace the battery with a 9-V battery (NEDA A1604, 6F22, or 006P).  
XWWarning  
To avoid false readings, which could lead to possible electric  
shock or personal injury, replace the battery as soon as the  
battery indicator (b) appears. If the display shows bAttthe  
Meter will not function until the battery is replaced.  
Replace the battery as follows, refer to Figure 1:  
1. Turn the rotary knob to OFF and remove the test leads from the terminals.  
2. Remove the battery door by using a standard-blade screwdriver to turn the battery  
door screws one-quarter turn counterclockwise.  
3. Remove the old battery and replace it with a new one.  
4. Align the battery leads so that they not pinched between the battery door and the case  
bottom.  
5. Secure the door by turning the screws one-quarter turn clockwise.  
Testing Fuses and Current Circuitry  
If a test lead is plugged into the mA/µA or A terminal and the rotary knob is turned to a  
non-current function, the Meter chirps and flashes “LEAd” if the fuse associated with that  
current terminal is good. If the Meter does not chirp or flash “LEAd”, the fuse is bad and  
must be replaced. Refer to Table 17 for the appropriate replacement fuse.  
After replacing the fuse, use the following procedure to verify the integrity of the new  
fuse and the current circuitry. Refer to Figure 3.  
1. Turn the rotary knob to N.  
2. To test F2, insert a test lead into the Iinput terminal and touch the probe to  
the A input terminal.  
Note  
The input receptacles contain split contacts. Be sure to touch the probe to  
the half of the receptacle nearest the LCD.  
3. The display should indicate between 00.0 and 00.5 . If the display reads OL,  
replace the fuse and test again. If the display reads another value, further servicing is  
required.  
4. To test F1, move the probe from the A input terminal to the mA/UA input terminal.  
5. The display should read between 0.995 kand 1.005 k. If the display reads OL,  
replace the fuse and test again. If the display reads another value, further servicing is  
required.  
XWWarning  
To avoid electrical shock or personal injury:  
Remove the test leads and any input signals before  
replacing the battery or fuses.  
Install ONLY specified replacement fuses with the  
amperage, voltage, and speed ratings shown in Table 17.  
12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Digital Multimeter  
Required Equipment  
TRUE RMS MULTIMETER  
87  
V
Good F2 fuse: 00.0 to  
00.5 Ω  
C/  
F
˚
˚
MIN MAX  
RANGE  
AutoHOLD  
Peak MIN MAX  
Replace fuse: OL  
REL  
Hz  
%
4½ DIGITS  
1
Second  
mV  
mA  
A
V
LO  
A
V
OFF  
Touch top half  
of input contacts  
A
mA  
A
COM  
V
400mA  
FUSED  
10A MAX  
FUSED  
TRUE RMS MULTIMETER  
87  
V
C/  
F
˚
˚
MIN MAX  
Peak MIN MAX  
RANGE  
AutoHOLD  
REL  
Hz  
%
4½ DIGITS  
1
Second  
mV  
mA  
A
Good F1 fuse: 0.995 kto  
1.005 kΩ  
V
LO  
A
V
OFF  
A
mA  
A
COM  
V
Replace fuse: OL  
400mA  
FUSED  
10A
FU
aom5f.eps  
Figure 3. Testing the Current Input Fuses  
Replacing the Fuses  
To replace the fuse(s), perform the following procedure.  
1. To open the Meter, refer to “Opening the Meter Case”. See Figure 1.  
2. Grasp the fuse in the center with needle nose pliers. Pull straight up on the fuse to  
remove it from the fuse clips.  
3. Install ONLY specified replacement fuses with the amperage, voltage, and speed  
ratings shown in Table 17.  
4. To close the Meter, refer to “Reassembling the Meter Case”.  
Required Equipment  
Required equipment for the performance tests is listed in Table 13. If the recommended  
models are not available, equipment with equivalent specifications may be used.  
XWWarning  
To avoid shock or injury, do not perform the verification  
tests or calibration adjustment procedures described in this  
manual unless you are qualified to do so.  
Repairs or servicing should be performed only by qualified  
personnel.  
13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
80 Series V  
Calibration Manual  
Table 13. Required Equipment  
Required Characteristics  
Equipment  
Calibrator  
Recommended Model  
AC Voltage Range: 0 - 1000 V ac  
Accuracy: ± 0.12 %  
Fluke 5520A Multi-Product  
Calibrator or equivalent  
Frequency Range: 60 - 20000 Hz  
Accuracy: ± 3 %  
DC Voltage Range: 0 - 1000 V dc  
Accuracy: ± 0.012 %  
Current Range: 350 µA - 2 A  
Accuracy: AC (60 Hz to 1 kHz): ± 0.25 %  
DC: ± 0.05 %  
Frequency Source: 19.999 kHz - 199.99 kHz  
Accuracy: ± 0.0025 %  
Amplitude: 150 mV to 6V rms  
Accuracy: ± 5 %  
Range: 1 - 100 MΩ  
Accuracy: 0.065 %  
Function Generator  
Frequency = 900 kHz  
Amplitude = 8.3V  
HP33120  
Burst mode = 1  
Burst rate = 100Hz  
Burst Phase = -90 degrees  
Fluke 80 AK  
K-type  
Fluke 80 AK  
TC Adapter Accessory  
K-type Thermocouple  
K-type, mini-plug on both ends  
Performance Tests  
The following performance tests verify the complete operability of the Meter and check  
the accuracy of each Meter function against the Meter’s specifications. Performance tests  
should be performed annually to ensure that the Meter is within accuracy specifications.  
Accuracy specifications are valid for a period of one year after calibration adjustment,  
when measured at an operating temperature of 18 °C to 28 °C and at a maximum of 90 %  
relative humidity.  
To perform the following tests, it is not necessary to open the case. No adjustments are  
necessary. Make the required connections, apply the designated inputs, and determine if  
the reading on the Meter display falls within the acceptable range indicated.  
Note  
If the Meter fails any of these tests, it needs calibration adjustment or repair.  
14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Digital Multimeter  
Performance Tests  
Basic Operability Tests  
Refer to the following sections to test the basic operability of the Meter.  
Testing the Fuses  
Refer to “Testing the Fuses”.  
Testing the Display  
Turn the Meter on while holding down Dto view all segments of the display.  
Compare the display with the appropriate examples in Figure 4 and Table 14.  
ayi04.eps  
Figure 4. Display Features  
Table 14. Display Features  
Number  
Feature  
Indication  
Y
Polarity indicator for the analog bar graph.  
Positive or negative slope indicator for Hz/duty cycle triggering.  
The continuity beeper is on.  
A
TrigY  
X
B
C
D
W
Relative (REL) mode is active.  
g
Smoothing is active.  
Indicates negative readings. In relative mode, this sign indicates that the  
present input is less than the stored reference.  
-
E
F
Indicates the presence of a high voltage input. Appears if the input voltage is  
30 V or greater (ac or dc). Also appears in low pass filter mode. Also appears  
in cal, Hz, and duty cycle modes.  
Z
RS  
S
G
H
I
J
K
AutoHOLD is active.  
Display Hold is active.  
p
Indicates the Meter is in Peak Min Max mode and the response time is 250 µs  
Indicators for minimum-maximum recording mode.  
m MAX  
MIN AVG  
K
Low pass filter mode.  
15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
80 Series V  
Calibration Manual  
Table 13. Display Features (cont.)  
Indication  
Number  
Feature  
The battery is low. XWWarning: To avoid false readings, which could lead to  
possible electric shock or personal injury, replace the battery as soon as the  
battery indicator appears.  
b
L
M
Amperes (amps), Microamp, Milliamp  
Volts, Millivolts  
A, µA, mA  
V, mV  
Microfarad, Nanofarad  
µF, nF  
nS  
Nanosiemens  
Percent. Used for duty cycle measurements.  
Ohm, Megohm, Kilohm  
%
e, Me, ke  
Hz, kHz  
AC DC  
°C, °F  
Hertz, Kilohertz  
Alternating current, direct current  
Degrees Celsius, Degrees Fahrenheit  
Displays selected range  
N
O
610000 mV  
The Meter is in high resolution  
(Hi Res) mode. HiRes=19,999  
P
HiRes  
The Meter is in autorange mode and automatically selects the range with the  
best resolution.  
Auto  
Q
R
The Meter is in manual range mode.  
Manual  
The number of segments is relative to the full-scale value of the selected  
range. In normal operation 0 (zero) is on the left. The polarity indicator at the  
left of the graph indicates the polarity of the input. The graph does not operate  
with the capacitance, frequency counter functions, temperature, or peak min  
max. For more information, see “Bar Graph”. The bar graph also has a zoom  
function, as described under "Zoom Mode".  
RPMon  
oconventional (4 cycle) Counts every other revolution.  
nwaste spark of 2 cycle. Counts every revolution.  
S
--  
Overload condition is detected.  
0L  
Error Messages  
bAtt  
Replace the battery immediately.  
diSC  
In the capacitance function, too much electrical charge is present on the capacitor being tested.  
EEPr  
Err  
Invalid EEPROM data. Have Meter serviced.  
Invalid calibration data. Calibrate Meter.  
CAL  
Err  
WTest lead alert. Displayed when the test leads are in the A or mA/µA terminal and the selected  
rotary switch position does not correspond to the terminal being used.  
LEAd  
16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Digital Multimeter  
Performance Tests  
Testing the Pushbuttons  
To test the pushbuttons  
1. Turn the Meter rotary knob to J.  
2. Press each button and note that the meter responds with a beep for each button  
press.  
3. Press and hold Ba second time to exit MIN MAX mode.  
Testing Meter Accuracy  
Perform the accuracy test steps in Table 15.  
Table 15. Accuracy Tests  
Display Reading  
83 and 88  
Test  
Step  
Range  
5500A Output  
Function  
87  
1
600 mV  
600 mV  
6 V  
330 mV, 60 Hz  
327.9 to 332.1  
N/A  
327.3 to 332.7  
586.0 to 614.0  
3.275 to 3.325  
3.214 to 3.386  
32.75 to 33.25  
32.14 to 33.86  
327.5 to 332.5  
323.0 to 337.0  
494 to 506  
K
AC Volts  
2
600 mV, 13 kHz  
3.3 V, 60 Hz  
3
3.281 to 3.319  
N/A  
4
6 V  
3.3 V, 20 kHz  
33 V, 60 Hz  
5
60 V  
32.81 to 33.19  
N/A  
6
60 V  
33 V, 20 kHz  
7
600 V  
600 V  
1000 V  
1000 V  
600 mV  
600 mV  
330 V, 60 Hz  
328.1 to 331.9  
N/A  
8
330 V, 2.5 kHz  
500 V, 60 Hz  
9
495 to 505  
986 to 1014  
99.93 to 99.97  
199.48 to 199.52  
10  
11  
12  
1000 V, 1 kHz  
150 mV, 99.95 kHz  
150 mV, 199.50 kHz  
986 to 1014  
99.93 to 99.97  
199.48 to 199.52  
K
AC Volts  
Hz  
Frequency  
13  
14  
15  
Sensitivity  
6 V  
0.7 V, 99.95 kHz  
7 V, 99.95 kHz  
99.93 to 99.97  
99.93 to 99.97  
999.8 to 1000.2  
99.93 to 99.97  
99.93 to 99.97  
999.8 to 1000.2  
60 V  
6 V  
3.4 V, 1 kHz Sq. Wave  
VHz  
Trigger level  
16  
6 V  
5 V, 1 kHz, DC offset 2.5 V  
Sq. Wave  
49.7% to 50.3 %  
49.7 % to 50.3 %  
VHz  
Duty Cycle  
17  
18  
6V  
3.3 V dc  
33 V dc  
3.297 to 3.303  
32.97 to 33.03  
3.296 to 3.304  
32.96 to 33.04  
V
DC Volts  
60 V  
19  
20  
21  
22  
600 V  
330 V dc  
329.7 to 330.3  
998 to 1002  
329.6 to 330.4  
998 to 1002  
1000 V  
600 mV  
600 mV  
1000 V dc  
33 mV dc  
330 mV dc  
32.9 to 33.1  
32.8 to 33.2  
mV  
DC Volts  
329.6 to 330.4  
328.9 to 331.1  
17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
80 Series V  
Calibration Manual  
Table 14. Accuracy Tests (cont.)  
5500A Output  
Display Reading  
83 and 88  
Test  
Function  
Step  
Range  
87  
23  
24  
25  
26  
27  
28  
29  
30  
31  
e
600 Ω  
6 kΩ  
330 ( Use 2 wire Comp)1  
3.3 k(Use 2 wire Comp)1  
329.1 to 330.9  
3.292 to 3.308  
32.92 to 33.08  
327.9 to 332.1  
3.279 to 3.321  
29.67 to 30.33  
- 0.30 to 0.30  
9.60 to 10.40  
2.939 to 3.061  
328.5 to 331.5  
3.286 to 3.314  
32.86 to 33.14  
327.6 to 332.4  
3.276 to 3.324  
29.67 to 30.33  
- 0.30 to 0.30  
9.60 to 10.40  
2.939 to 3.061  
Ohms  
60 kΩ  
600 kΩ  
6 MΩ  
50 MΩ  
60 nS  
60 nS  
6 V  
33 kΩ  
330 kΩ  
3.3 MΩ  
30 MΩ  
nS  
Open input  
100 MΩ  
Conductance  
3.0 V dc  
G
Diode  
32  
6 A  
3.0 A, 60 Hz  
3.0 A  
2.962 to 3.038  
2.984 to 3.016  
9.94 to 10.06  
2.968 to 3.032  
2.990 to 3.010  
9.96 to 10.04  
?
AC Amps  
33  
6 A  
A
DC Amps  
10A6  
10A  
33B  
A
DC Amps  
34  
35  
36  
37  
38  
39  
60 mA  
33 mA, 60 Hz  
330 mA, 60 Hz  
33 mA  
32.58 to 33.42  
325.8 to 334.2  
32.83 to 33.17  
328.5 to 331.5  
325.8 to 334.2  
3258 to 3342  
32.65 to 33.35  
326.5 to 333.5  
32.89 to 33.11  
329.1 to 330.9  
326.5 to 333.5  
3265 to 3335  
?
m
400 mA  
60 mA  
AC Milliamps  
m[  
DC Milliamp  
400 mA  
600 µA  
6000 µA  
330 mA  
330 µA, 60 Hz  
3300 µA, 60 Hz  
?
AC  
U
Microamps  
40  
41  
U[  
600 µA  
330 µA  
328.3 to 331.7  
3285 to 3315  
328.9 to 331.1  
3291 to 3309  
DC  
Microamps  
6000 µA  
3300 µA  
42  
43  
44  
45  
46  
E
10 nf  
Open input2  
5 nf5  
0.21 to 0.31  
04.7 to 05.3  
09.2 to 09.8  
N/A  
0.21 to 0.31  
04.7 to 05.3  
09.2 to 09.8  
376 to 408  
226 to 3404  
Capacitance  
100 nf  
100 µf  
1000 V  
1000 V  
9.5 µf  
400 V, 400 Hz  
400 V, 800 Hz4  
K
Low Pass  
N/A  
Filter  
18  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Digital Multimeter  
Performance Tests  
Table 14. Accuracy Tests (cont.)  
5500A Output  
Display Reading  
83 and 88  
Test  
Function  
Step  
Range  
87  
6 V dc  
8 Vpp, 2 kHz Sq.  
Wave, DC offset 2 V  
Max = 5.896 to 6.104  
Min = -1.898 to -2.102  
47  
48  
L
(87 and 88 only)  
Peak Min/Max  
0 °C  
-1.0 to 1.0  
98.0 to 102.0  
-1.0 to 1.0  
49  
50  
L
m
(87 and 88 only)  
Temperature 3  
100 °C  
98.0 to 102.0  
Backlight  
Press backlight button  
Press backlight button  
Press backlight button  
Backlight comes on  
51  
52  
53  
Backlight Intensifies  
Backlight off  
1. Or short test leads and use REL to offset test lead resistance.  
2. Remove test leads from unit.  
3. To ensure accurate measurement, the Meter and thermocouple adapter must be at the same temperature. After  
connecting the thermocouple adapter to the Meter allow for reading to stabalize before recording display reading.  
4. The Meter accuracy is not specified at this input signal frequency with Low-pass filter selected. The display reading  
shown, check that the Low-pass filter is active and follows an expected roll-off curve.  
5. Use REL to compensate for internal Meter and lead capacitance. Test leads must be disconnected from the  
calibrator before using REL.  
6. W10 A continuous up to 35 °C; < 20 minutes on, 5 minutes off at 35 °C to 55 °C. 20 A for 30 seconds maximum;  
> 10 A unspecified.  
Testing the Inductive Pickup (88 Only)  
To test the inductive pickup, a function generator output will simulate automobile spark  
plug signals on a loop of wire containing a 10 eresistor. The pickup will be clamped to  
the wire and output voltage from the pickup will be monitored by an oscilloscope.  
Perform the following procedure to test the inductive pickup:  
1. Solder a 10-inch piece of 14 AWG wire to one end of a 10 e1 % resistor.  
2. Connect the other end of the 10 eresistor to the terminal LOW output of the  
function generator. Place the other end of the 14-AWG wire to the HIGH output of  
the function generator. See Figure 5.  
3. Connect a 10X scope probe from channel 2 (dc-coupled) of the oscilloscope across  
the 10 eresistor.  
4. Clamp the inductive pickup to the wire loop on the HIGH side of the resistor as  
shown in Figure 5.  
Make sure that the jaws of the inductive pickup are closed completely, and that the  
side of the inductive pickup that says "SPARK PLUG SIDE" points toward the  
HIGH output of the function generator.  
5. Connect a 10X scope probe from channel 1 (dc-coupled) of the oscilloscope across  
the output of the inductive pickup.  
19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
80 Series V  
Calibration Manual  
6. Set up the function generator as follows:  
Frequency  
Amplitude  
Burst Mode  
Burst Count  
Burst Rate  
Burst Phase  
900 kHz  
8.3 V  
yes  
1
100 Hz  
-90 Degrees  
7. Set the oscilloscope for 0.5 V/DIV @ 0.5 µs/DIV.  
8. Trigger the waveform on channel 2.  
9. Adjust the amplitude of the function generator to produce a 3 VP-P triangle wave.  
See Figure 6.  
10. Set the oscilloscope for 1.0 V/DIV @ 5.0 ms/DIV.  
11. Trigger the waveform on Channel 1.  
12. Check that the peak voltage is greater than 5.7 V and decays to less than 1.0 V  
between pulses. See Figure 6. Record the peak value for later use.  
13. Adjust the function generator output so the peak voltage is 6 V.  
14. Set the scope for 1.0 ms/DIV and trigger waveform.  
15. Check the amplitude after 3.0 ms from the waveform peak, the voltage amplitude is  
2.4 V +0.5/-0.8 (1.6 V to 2.9 V). (See Figure 3-5(C).)  
16. Re-adjust the function generator output to obtain the value recorded in step 12.  
17. Set the scope for 5.0 ms/DIV.  
18. Turn the inductive pickup so that "SPARK PLUG SIDE" points along the wire  
connected to the LOW output of the function generator. Check that the waveform is  
less than 2 V.  
20  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Digital Multimeter  
Performance Tests  
FUNCTION GENERATOR  
OSCILLOSCOPE  
10  
CH1  
CH2  
OUTPUT  
RESISTOR  
RED  
WIRE LOOP  
(14 AWG)  
SPARK PLUG SIDEINDUCTIVE  
PICKUP  
aad08f.eps  
Figure 5. Setup for Inductive Pickup Test  
21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
80 Series V  
Calibration Manual  
aad09f.wmf  
Figure 6. Waveform for Inductive Pickup Test  
22  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Digital Multimeter  
Calibration Adjustment  
Calibration Adjustment  
The Meter features closed-case calibration adjustment using known reference sources.  
The Meter measures the applied reference source, calculates correction factors and stores  
the correction factors in nonvolatile memory.  
The following sections present the features and Meter pushbutton functions that can be  
used during the Calibration Adjustment Procedure. Perform the Calibration Adjustment  
Procedure should the Meter fail any performance test listed in Table 15.  
Calibration Adjustment Counter  
The Meter contains a calibration adjustment counter. The counter is incremented each  
time a Calibration Adjustment Procedure is completed. The value in the counter can be  
recorded and used to show that no adjustments have been made during a calibration  
cycle.  
Use the following steps to view the Meter's calibration counter.  
1. While holding down B, turn the rotary knob from OFF to VAC. The Meter  
should display “ZCAL”.  
2. Press Donce to see the calibration counter. For example "n001".  
3. Turn the rotary knob to OFF.  
Calibration Adjustment Password  
To start the Calibration Adjustment Procedure, the correct 4-button password must be  
entered. The password can be changed or reset to the default as described in following  
paragraphs. The default password is “1234”.  
Changing the Password  
Use the following steps to change the Meter's password:  
1. While holding down B, turn the rotary knob from OFF to VAC. The Meter  
displays “ZCAL”.  
2. Press Donce to see the calibration counter.  
3. Press Dagain to start the password entry. The Meter displays "????".  
4. The Meter buttons represent the digit indicated below when entering or changing the  
password:  
A= 1 B= 2 C= 3 D= 4  
H= 5  
E= 6 F= 7 G= 8  
Press the 4 buttons to enter the old password. If changing the password for the first  
time, enter A(1) B(2) C(3) D(4).  
5. Press Cto change the password. The Meter displays "----" if the old password is  
correct. If the password is not correct, the Meter emits a double beep, displays "????"  
and the password must be entered again. Repeat step 4.  
6. Press the 4 buttons of the new password.  
7. Press Dto store the new password.  
23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
80 Series V  
Calibration Manual  
Restoring the Default Password  
If the calibration password is forgotten, the default password (1234) can be restored using  
the following steps.  
1. While holding down B, turn the rotary knob from OFF to VAC. The Meter  
displays “ZCAL”.  
2. Remove the Meter's top case. Leave the PCA in the bottom case. (See “Opening the  
Meter Case”.)  
XWWarning  
To avoid electrical shock or personal injury, remove the test  
leads and any input signal before removing the Meter's top  
case.  
3. Through an access hole provided in the top shield, short across the keypads on the  
PCA. See Figure 7. The Meter should beep. The default password is now restored.  
4. Replace the Meter's top case and turn the rotary knob to OFF. (See “Reassembling  
the Meter Case).  
Access Hole For  
PCA Keypad  
ama01f.eps  
Figure 7. Restoring the Default Password  
24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Digital Multimeter  
Meter Buttons Used in the Calibration Steps  
The Meter buttons behave as follows when performing the Calibration Adjustment  
Procedure. This may be of help determining why a calibration step is not accepted and for  
determining the input value without referring to Table 16.  
Press and hold to show the measured value. The measurement value is not  
A
calibrated so it may not match the input value. This is normal.  
B
G
Press and hold to display the required input amplitude.  
Press and hold to display the frequency of the required input.  
Press to store the calibration value and advance to the next step. This button  
is also used to exit calibration mode after the calibration adjustment  
sequence is complete.  
D
Calibration Adjustment Procedure  
Use the following steps to adjust the Meter's calibration. If the Meter is turned off before  
completion of the adjustment procedure, the calibration constants are not changed.  
1. While holding down B, turn the rotary knob from OFF to VAC. The Meter  
displays “ZCAL”.  
2. Press Donce to see the calibration counter.  
3. Press Dagain to start the password entry. The Meter displays "????".  
4. Press 4 buttons to enter the password.  
5. Press Dto go to the first calibration step. The Meter displays "C-01" if the  
password is correct. If the password is not correct, the Meter emits a double beep,  
displays "????" and the password must be entered again. Repeat step 4.  
6. Using Table 16, apply the input value listed for each calibration adjustment step. For  
each step, position the rotary switch and apply the input to the terminals as indicated  
in the table.  
7. After each input value is applied, press Dto accept the value and proceed to the  
next step (C-02 and so forth).  
Notes  
After pressing D, wait until the step number advances before changing  
the calibrator source or turning the Meter rotary knob.  
If the Meter rotary knob is not in the correct position, or if the measured  
value is not within the anticipated range of the input value, the Meter emits  
a double beep and will not continue to the next step.  
Some adjustment steps take longer to execute than others (10 to 15  
seconds). For these steps, the Meter will beep when the step is complete.  
Not all steps have this feature.  
8. After the final step, the display shows "End" to indicate that the calibration  
adjustment is complete. Press Dto go to meter mode.  
25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
80 Series V  
Calibration Manual  
Notes  
Set the calibrator to Standby prior to changing the function switch position  
and or after completing adjustment of each function.  
If the calibration adjustment procedure is not completed correctly, the  
Meter will not operate correctly.  
Table 16. Calibration Adjustment Steps  
Input Value  
Function  
(Switch Position)  
Adjustment  
Step  
Input Terminal  
Fluke 83-V and  
88-V  
Fluke 87-V  
C-01  
C-02  
C-03  
C-04  
C-05  
C-06  
C-07  
C-08  
C-09  
C-10  
C-11  
C-12  
C-13  
C-14  
C-15  
C-16  
C-17  
C-18  
C-19  
C-20  
600.0 mV, 60 Hz  
600.0 mV, 5 kHz  
6.000 V, 60 Hz  
6.000 V, 5 kHz  
60.00 V, 60 Hz  
60.00 V, 5 kHz  
600.0 V, 60 Hz  
600.0 V, 5 kHz  
6.000 V  
600.0 mV, 60 Hz  
600.0 mV, 20 kHz  
6.000 V, 60 Hz  
6.000 V, 20 kHz  
60.00 V, 60 Hz  
60.00 V, 20 kHz  
600.0 V, 60 Hz  
600.0 V, 10 kHz  
6.000 V  
K
I
(AC Volts)  
L
(DC Volts)  
60.00 V  
60.00 V  
600.0 V  
600.0 V  
600.0 mV  
600.0 mV  
L
m
(DC Millivolts)  
60.00 mV  
60.00 mV  
600.0  
600.0  
e
(Ohms)  
6.000 k  
6.000 k  
60.00 k  
60.00 k  
600.0 k  
600.0 k  
6.000 M  
6.000 M  
0.000  
0.000  
50.0 M  
50.0 M  
O
C-21  
3.000 V  
3.000 V  
(Diode Test)  
C-22  
C-23  
6.000 A, 60 Hz  
6.000 A dc  
6.000 A, 60 Hz  
6.000 A dc  
A
A
(Amps)  
C-24  
C-25  
C-26  
C-27  
60.00 mA, 60 Hz  
400.0 mA, 60 Hz  
60.00 mA dc  
60.00 mA, 60 Hz  
400.0 mA, 60 Hz  
60.00 mA dc  
mA  
(Milliamps)  
mA /UA  
400.0 mA dc  
400.0 mA dc  
600.0 µA ac,  
60 Hz  
µA  
C-28  
600.0 µA ac, 60 Hz  
(Microamps)  
C-29  
C-30  
C-31  
6000 µA, 60 Hz  
600.0 µA dc  
6000 µA dc  
6000 µA, 60 Hz  
600.0 µA dc  
6000 µA dc  
26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Digital Multimeter  
Service and Parts  
Service and Parts  
Replacement parts are shown in Table 17 and Figure 8. To order parts and accessories,  
refer to “Contacting Fluke”.  
Table 17. 80 Series V Final Assembly  
Description  
Part  
Number  
Ref Des  
Qty  
AC72  
AC72  
BT1  
Alligator Clip, Black  
Alligator Clip, Red  
Battery, 9 V  
1670652  
1670641  
2139179  
2064217  
2074057  
943121  
803293  
832246  
448456  
817460  
674853  
824466  
1
1
1
1
1
1
1
3
5
2
1
2
BT2  
Cable Assy, 9 V Battery Snap  
Lightpipe  
CR6  
F1W  
F2W  
H2-4  
H5-9  
J1-2  
J3  
Fuse, 0.440 A, 1000 V, FAST  
Fuse, 11 A, 1000 V, FAST  
Screw, Case  
Screw, Bottom Shield  
Elastomeric Connector  
Top Shield Contact  
Foot, Non-Skid  
MP10,  
MP11  
MP2  
MP4  
MP5  
MP5  
MP5  
MP6  
MP8  
MP9  
MP13  
MP14  
MP15  
MP22  
Shield, Top  
2073906  
2074025  
2074002  
2073992  
2115202  
2073871  
2100482  
822643  
1
1
Shield, Bottom  
Case Top (PAD XFER) With Window (83-5)  
Case Top (PAD XFER) with Window (87-5)  
Case Top (PAD XFER) with Window (88-5)  
Case Bottom  
1
1
1
1
1
1
1
1
1
4
Knob, Switch (PAD XFER)  
Detent, Knob  
Shock Absorber  
828541  
O-Ring, Input Receptacle  
Holster w/ Tilt Stand  
831933  
2074033  
2073938  
1567683  
Battery Door  
MP27-  
MP30  
Contact RSOB  
MP31  
MP31  
MP31  
MP41  
Mask, LCD (PAD XFER) (83-5)  
Mask, LCD (PAD XFER) (87-5)  
Mask, LCD (PAD XFER) (88-5)  
Housing, RSOB  
2073961  
2073950  
2112410  
2073945  
948609  
1
1
1
1
2
MP390-  
391  
Access Door Fastener  
NA  
Tiltstand  
2074040  
2105884  
855742  
1
1
1
1
1
1
1
1
1
S2  
Keypad  
TL75  
TM1  
TM2  
TM 2  
TM3  
TM3  
Test Lead Set  
80 Series V Getting Started Manual (Multi-language)  
80 Series V Quick Reference Card  
88 V Quick Reference Card  
2101973  
2101986  
2279006  
2101999  
2278999  
2102915  
CD ROM (Contains 80 Series V Users Manual )  
CD ROM (Contains 88 V Users Manual )  
80 Series V Calibration Manual (this manual)  
TM4  
(not  
shown)  
U5  
LCD, 4.5 DIGIT,TN, Transflective, Bar Graph, OSPR80  
2065213  
1273113  
1
1
MP81  
80BK Thermocouple Assembly, K-Type, Beaded, Molded Dual Banana Plug,  
Coiled  
WTo ensure safety, use exact replacement only.  
27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
80 Series V  
Calibration Manual  
TL75  
Test Lead Set  
MP66  
MP8  
MP31  
U5  
MP5  
J1-2  
CR6  
MP2  
S2  
MP9  
Alligator Clips  
AC72  
MP13  
MP41  
J3  
MP81  
F2  
MP27-30  
F1  
W
MP4  
Holster  
MP15  
MP14  
BT2  
H5-9 (5)  
MP6  
H2-4 (3)  
Tilt Stand  
MP10-11  
TM1  
BT1  
TM3  
MP22  
TM2  
MP390-391  
aom015c.eps  
Figure 8. 80 Series V Final Assembly  
28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 

Electrolux Refrigerator TR926 User Manual
Estate Clothes Dryer EED4300V User Manual
Eton Weather Radio ESP2100 User Manual
Euro Pro Convection Oven TO320 User Manual
Everpure Water System A 20 20 User Manual
Excalibur electronic Games VR15 User Manual
Franke Consumer Products Ventilation Hood FBI 362 XS IB User Manual
Fujitsu Laptop T904 User Manual
Fujitsu Network Router P2X0 0202 01EN User Manual
Gaggenau Freezer RB 472 User Manual