Introduction  
The Model 80T-150UA Universal Temperature Probe is a self-contained  
temperature-to-voltage converter. The probe is designed to provide a  
direct temperature reading when it is connected to any high impedance  
DMM that is capable of 1 mV resolution and at least a 300-count full  
scale readout capability. Output is 1 mV per degree (Celsius or  
Fahrenheit). Two switch-selected temperature output scalings are  
provided: -50 to +150 °C or -58 to +302 °F. The probe will stand off 60 V  
dc or 30 V ac rms (42.4 V peak).  
The unit is housed in two separate but attached assemblies: a  
temperature probe and a temperature-to-voltage converter. The probe  
contains the temperature-sensing element and is electrically connected to  
the temperature-to-voltage converter through a 46-inch cable. A three-  
position switch on the converter acts as a power switch and is used for  
selecting Celsius or Fahrenheit scaling for the output. Two banana plugs  
with standard 0.75-inch spacing are provided for connecting the 80T-  
150UA to the DMM.  
Operating power for the 80T-150UA is derived from a standard 9 V  
battery. Typically, an alkaline battery provides more than 600 hours of  
continuous operation before replacement is necessary. An OFF switch is  
provided on the temperature-to-voltage converter to allow battery  
conservation when the unit is not in use. In addition, the OFF position of  
the power switch allows the battery condition to be determined via the  
external DMM.  
Temperature is measured by exposing the probe tip directly to the  
material to be measured (non-corrosive liquid, gas, or solid). A direct  
temperature reading is displayed on the DMM.  
Symbols  
XWarning: Risk of electric shock. Refer to the manual for  
information.  
W Attention: Refer to the manual for information about this  
feature.  
PN 3208738  
February 2008  
© 2008 Fluke Corporation. All rights reserved.  
All product names are trademarks of their respective companies.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
General  
Weight: 150 g (5.3 oz)  
Overall Length: 53.8 inch (1.36 m)  
Battery: Standard 9 V battery (NEDA #1604,6F22,006P)  
Battery Life: 600+ hours, typical (Alkaline Battery), 6.5 V minimum  
Output Termination: Standard 0.75-inch spaced double banana plug  
Probe Material: Glass-filled valox  
Probe Size: 0.6 inch maximum diameter  
Tip Material: Nickle Plated Copper  
Tip Size: 0.31 cm (0.12 inch diameter)  
Operating Notes  
The following paragraphs are intended to familiarize the operator with  
the 80T-150UA. The operator should read these paragraphs before  
attempting to operate the probe.  
Probe Limitations  
The 80T-150UA probe is constructed of a highly durable plastic and is  
suitable for measuring the temperature of liquids, gases, and solid  
surfaces up to 150 °C. When measuring temperature, observe the  
following precautions to prevent damage to the probe:  
1. Do not expose the probe end (probe tip plus about 2 inches of the  
probe body) to temperatures exceeding +150 °C (+302 °F). The  
remainder of the probe body should not be exposed to temperatures  
above +70 °C (+160 °F).  
2. Do not expose the probe end (probe tip plus about 2 inches of the  
probe body) to temperatures below -50 °C (-58 °F). The remainder  
of the probe body should not be exposed to temperatures below  
-40 °C (-40 °F).  
3. For liquid measurements, recommended applications range from  
water, lubricants, and fuels to most solvents. Liquids as shallow as  
1.27 cm (0.5 inch) can be measured since the temperature sensor is  
inch the probe tip.  
XWWarning  
To avoid electrical shock, do not use this instrument when  
voltages exceeding 60 V dc or 30 V ac rms (42.4 V peak )  
are present. The probe tip is electrically connected to the  
output terminals.  
WCaution  
Long-term exposure of the probe to corrosive  
environments will result in pitting and deterioration of the  
aluminum probe tip. If equipment is used in a manner not  
specified by the manufacturer, the protection provided by  
the equipment may be impaired.  
Error Sources  
When the probe tip is applied to a solid surface, it draws or sinks heat  
from the surface. Therefore, if the measured surface has a low mass  
(e.g., a transistor case), the indicated temperature may be lower than  
the actual temperature.  
Similarly, a steady-state error or gradient exists between the measured  
surface and the sensing device in the probe tip. This is due to the flow of  
heat from the measurement surface to the probe body. The effect of the  
steady-state error increases as the differential between ambient and  
surface temperature increases.  
To determine the actual surface temperature of a device, both the heat-  
sinking and steady-state errors must be considered. The correction curve  
given in Figure 1 approximates the effect of both error sources on TO-3,  
TO-5, and TO-18 transistor cases.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F1.eps  
Figure 1. Initial Case Temperature Above Ambient vs Meter  
Reading Above Ambient  
Operation  
Use the following procedure to operate the 80T-150UA probe:  
1. Connect the banana plugs on the 80T-150UA to the input terminals  
of a high impedance DMM. Observe polarity.  
2. Select a dc voltage range that will provide at least 1 mV resolution  
(1 mV/degree) and a full scale readout that will encompass the  
expected temperature. The 2 V range of a  
3 1/2-digit DMM is adequate. Ignore readings of less than 1° when  
a more sensitive DMM is used.  
3. Set the 80T-150UA power switch to °C or °F, and energize the  
DMM.  
4. Firmly touch the probe tip to the surface to be measured, or expose  
it to a liquid or gas. The DMM will display the temperature in  
degrees. Vary the probe angle and pressure when measuring solid  
surface temperatures. The highest stabilized reading will be the  
most accurate. (See the following measuring technique.)  
WCaution  
The force exerted on the probe tip should not exceed 15  
pounds.  
Measuring Technique  
Here are some suggestions for improving the accuracy of your  
temperature measurements:  
1. When measuring higher than ambient temperatures, adjust the  
connection between the probe and the surface until you get the  
highest temperature reading.  
2. When measuring lower than ambient temperatures, adjust the  
connection between the probe and the surface until you get the  
lowest temperature reading.  
3. When measuring near ambient temperatures, make the reading  
when the multimeter readout is most stable.  
General Maintenance  
Access Information  
The battery and the calibration pots are located on the interior of the  
temperature-to-voltage converter assembly. Access to these locations is  
accomplished by removing the screw from the bottom side of the  
assembly and removing the top of the plastic box.  
Battery Condition Test  
1. Set the power switch to the OFF position.  
2. Connect the 80T-150UA to the DMM.  
3. Set the DMM to the 200 or 400 mV dc range.  
4. Read the battery test voltage on the DMM or test instrument. For  
DMMs with 10 MΩ input impedance, a minimum reading of 100 mV  
is acceptable. For test instruments with 1 MΩ input impedances,  
such as ScopeMeter®, a minimum reading of 75 mV is acceptable,  
With these minimum readings, approximately 100 hours of battery  
life remain.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Battery Replacement  
XWWarning  
To avoid electrical shock, remove the probe from the  
measurement surface before opening the case. Totally  
reassemble the instrument before attempting to use it.  
1. Set the power switch to the OFF position.  
2. Disconnect the 80T-150UA from the DMM.  
3. Turn the 80T-150UA so the power switch is facing down. Remove  
the single screw located between the banana plugs.  
4. Grasp one case half in each hand. Pull the two halves apart,  
beginning at the end with the banana plugs.  
5. Remove and replace the battery.  
6. Reassemble the 80T-150UA as follows. Mate the two case halves at  
the end where the cable enters the case, then "hinge" the two  
halves together. Replace the case screw, being careful not to pinch  
the probe cable or battery wires.  
Performance Test  
Complete the calibration procedure without opening the temperature-to-  
voltage converter assembly and without making any calibration  
adjustments. Observe the readings given in [brackets]. Other readings  
are for calibration only.  
Calibration  
The calibration cycle of one year is recommended to maintain the unit  
within the specifications given earlier. The equipment required for  
calibration is listed in the table following the calibration procedure.  
Note  
Values given in brackets apply to the Performance Test.  
Perform the following steps to calibrate the 80T-150UA:  
1. Access the interior of the temperature-to-voltage converter by  
removing the bottom case screw and separating the case halves.  
2. Connect the 80T-150UA to a DVM with 10 μV resolution, and select  
mV dc range.  
3. Select the °C position of the switch. Immerse the probe tip 5.1 cm (2  
inches) into a mercury thermometer monitored 0 °C bath, and allow  
a full 5 minutes for the reading to stabilize.  
4. Adjust VR3 and VR1 (see Figure 2) to obtain the following reading:  
0.00 0.05 mV dc [0 0.2 mV dc]  
5. Select the °F position of the switch, and adjust VR2 to obtain the  
following reading:  
32.0 0.1 mV dc [32.0 0.4 mV dc]  
6. Select the °C position, and move the probe tip to a 70 °C bath and  
again allow the reading to stabilize.  
7. Adjust VR4 to obtain a DVM reading that agrees with the bath  
temperature (BT) as monitored by a mercury thermometer.  
°C BT 0.05 mV dc [BT 0.2 mV dc]  
8. Select the °F position, and verify that the output is within 4 mV dc  
of the bath thermometer reading. If necessary, change the DVM  
range to obtain an on-scale reading.  
9. Return the probe tip to the 0 °C bath wait 10 minutes, then check the  
output. If readjustment is necessary, repeat steps 4 through 8 until  
the readings can be obtained without adjustment.  
10. Set the 80T-150UA switch to the OFF position, and remove the 80T-  
150UA from the DVM.  
11. Reassemble the 80T-150UA.  
12. The 80T-150UA is now calibrated.  
Cleaning  
Clean the test tool with a damp cloth and a mild soap. Do not use  
abrasives, solvents, or alcohol. These may damage the text on the test  
tool.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Sensing Probe and Cable Assembly  
VR2  
VR4 VR3 VR1  
MP101  
MP102  
F4.eps  
Figure 2. 80T-150UA Temperature Probe  
Test Equipment Requirements  
Minimum Use  
Specifications  
Recommended  
Model  
Instrument Type  
Mercury Thermometer  
0.1 °C Resolution  
Princo Model SAMA-  
CP45  
Dewar Flask and Cap  
1-Pint Capacity (for Ice Bath)  
1-Pint Capacity  
Thermos Bottle  
Metal or Glass Container  
Suitable for Boiling  
Water  
Digital Voltmeter  
100 mV Range with 10 μV  
Fluke Model 8845A  
Resolution  
1000 mV Range with 100 μV  
To contact Fluke or for service, call one of the following telephone numbers:  
USA: 1-888-44-FLUKE (1-888-443-5853), Canada: 1-800-36-FLUKE (1-800-363-  
5853), Europe: +31 402-675-200, Japan: +81-3-3434-0181, Singapore: +65-738-  
5655, Anywhere in the world: +1-425-446-5500  
Or, visit Fluke's Web site at www.fluke.com.  
Fluke Corporation  
P.O. Box 9090  
Everett WA 98206-9090  
USA  
Fluke Europe B.V.  
P.O. Box 1186  
5602 B.D. Eindhoven  
The Netherlands  
LIMITED WARRANTY & LIMITATION OF LIABILITY  
Each Fluke product is warranted to be free from defects in material and workmanship under  
normal use and service. The warranty period is one year and begins on the date of shipment.  
Parts, product repairs and services are warranted for 90 days. This warranty extends only to the  
original buyer or end-user customer of a Fluke authorized reseller, and does not apply to fuses,  
disposable batteries or to any product which, in Fluke's opinion, has been misused, altered,  
neglected or damaged by accident or abnormal conditions of operation or handling. Fluke  
warrants that software will operate substantially in accordance with its functional specifications for  
90 days and that it has been properly recorded on non-defective media. Fluke does not warrant  
that software will be error free or operate without interruption.  
Fluke authorized resellers shall extend this warranty on new and unused products to end-user  
customers only but have no authority to extend a greater or different warranty on behalf of Fluke.  
Warranty support is available if product is purchased through a Fluke authorized sales outlet or  
Buyer has paid the applicable international price. Fluke reserves the right to invoice Buyer for  
importation costs of repair/replacement parts when product purchased in one country is submitted  
for repair in another country.  
Fluke's warranty obligation is limited, at Fluke's option, to refund of the purchase price, free of  
charge repair, or replacement of a defective product which is returned to a Fluke authorized  
service center within the warranty period.  
To obtain warranty service, contact your nearest Fluke authorized service center or send the  
product, with a description of the difficulty, postage and insurance prepaid (FOB Destination), to  
the nearest Fluke authorized service center. Fluke assumes no risk for damage in transit.  
Following warranty repair, the product will be returned to Buyer, transportation prepaid (FOB  
Destination). If Fluke determines that the failure was caused by misuse, alteration, accident or  
abnormal condition of operation or handling, Fluke will provide an estimate of repair costs and  
obtain authorization before commencing the work. Following repair, the product will be returned to  
the Buyer transportation prepaid and the Buyer will be billed for the repair and return  
transportation charges (FOB Shipping Point).  
THIS WARRANTY IS BUYER'S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL  
OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY  
IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.  
FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR  
CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, WHETHER  
ARISING FROM BREACH OF WARRANTY OR BASED ON CONTRACT, TORT, RELIANCE OR  
ANY OTHER THEORY.  
Since some countries or states do not allow limitation of the term of an implied warranty, or  
exclusion or limitation of incidental or consequential damages, the limitations and exclusions of  
this warranty may not apply to every buyer. If any provision of this Warranty is held invalid or  
unenforceable by a court of competent jurisdiction, such holding will not affect the validity or  
enforceability of any other provision.  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Echo Blower PB 770H User Manual
Excalibur electronic Home Safety Product H623 User Manual
Exmark Lawn Mower Turf Tracer Hydro User Manual
Fisher Price Motorized Toy Car N9733 User Manual
Fishman Music Pedal Archtop Guitar Pickup User Manual
FujiFilm Printer 40002 User Manual
Fujitsu Network Card LQFP 64P User Manual
Gateway Laptop TM 5150 User Manual
GBC Binding Machine 250T User Manual
Gefen Network Card GTV DIGAUD 2 AAUD User Manual