EXFO Photonic Solutions Div Power Supply IQS 1500 User Manual |
C A LIB R ATIO N P O WE R M E TE R
IQ S -1 5 0 0
15 0 0
R &D AN D MAN U FAC TU R IN G
Operates in the WDM window
Enables on-site power-meter calibration
Linearity of ± 0.01 dB
5 mm cooled detector
Telecommunications Test and Measurement
IQ S -1 5 0 0
C a lib ra tio n Mo d e
The IQS-1500’s Windows-compatible, highly intuitive GUI allows easy control via mouse, touchscreen or keyboard. Also, it lets you save multiple
user configurations and access the instrument’s online user guide.
A Start
Allows the user to select the calibration mode, start a new calibration
sequence with template file or view a report file.
A
B Information
B
C
D
Allows the user to enter general and specific information about current
E
F
measurement conditions. This information will be included in the
calibration summary.
G
H
C General
I
J
Contains user and module information currently used to perform
calibration.
K
L
D Conditions
Contains all the information pertaining to the conditions at which the
calibration is performed: source wavelength, power and bandwidth,
temperature and humidity.
E Device under Test (DUT)
Contains all the information pertaining to the DUT under calibration.
J DUT Measurements
F Uncertainties
Allows the user to enter DUT power measurements.
Allows the user to enter specific information pertaining to the
calculation of the calibration uncertainty.
K Summary
G Null
Displays complete calibration results, including mean power, calibration
factor and total uncertainty.
Allows the user to eliminate electronic offsets and dark currents to
ensure accuracy.
L Detailed calibration reports
H Calibration
Generates detailed calibration reports that include instrumentation
identification, environmental information, reference conditions and
a calibration summary for each wavelength.
Allows the user to perform calibration measurements and calculation.
I Ref. Meter Measurement
This auto-save function allows the user to select the number of
measurements and the intervals between readings for automated
measurements on the reference power meter (IQS-1500).
The IQS-1500’s Summary Window
Calibration factor for DUT
readings (in W)
1
Source wavelength
5
Wavelength setting of
power meter under test
2
Standard deviation of the
calibration factor
6
7
1
2
3
4
5
6
7
Mean power measured
by the IQS-1500
3
4
Expanded uncertainty of
corrected DUT readings
Number of calibration
measurements
Download from Www.Somanuals.com. All Manuals Search And Download.
IQ S -1 5 0 0
S P E C IF IC ATIO N S 1
Model
Detector type
Detector size (mm)
Power range1 (dBm)
Wavelength range (nm)
Linearity1,2 (dB)
IQS-1502-Q0
Ge
5
+5 to —60
750 to 1800
± 0.01
(+5 dBm to —50 dBm)
± 2 %3
IQS-1502-Q1
Ge
5
+5 to —60
750 to 1800
± 0.01
(+5 dBm to —50 dBm)
≤ ± 0.9 %4 at 3 λ
(—10 dBm CW)
STANDARD ACCES S ORIES
User guide, fiber-optic adapter
(FOA), one reference test jumper
and Certificate of Calibration
Notes
1. At 1310 nm and 1550 nm.
2. At a ± 1 °C constant temperature.
3. Q0 option, uncertainty at EXFO
reference conditions:
Uncertainty6
(at 1310.0 ± 0.1 nm and 1550.0 ± 0.1 nm)
(0 dBm to —10 dBm CW)
• ± 2 % uncertainty at (1310.0 ± 0.1) nm
and (1550.0 ± 0.1) nm, with 9/125 µm
(B) fiber, and source spectral width
(FWHM) < 10 nm.
• ± 3 % uncertainty from 750 nm to
1700 nm, with 62.5/125 µm (D) fiber,
and source spectral width (FWHM)
≤ 12 nm.
± 3 %3
(750 nm to 1700 nm)
(—30 dBm to —40 dBm CW)
0.001
± 3 %4
(750 nm to 1700 nm)
(—30 dBm to —40 dBm CW)
0.001
Power resolution (dB)
Wavelength resolution (nm)
Applicable fiber type (µm)
Connector type5
0.1
0.1
9/125 (B); 50/125 (C); 62.5/125 (D)
B: 89
9/125 (B); 50/125 (C); 62.5/125 (D)
B: 58; 88; 89; 90; 91
C/D: 50; 54; 74
• FC/UPC connector (ceramic ferrule)
with FOA-322.
GENERAL S PECIFICATIONS
4. Q1 option, calibration at NIST
a) • ≤ ± 0.9 % uncertainty at three
user-specified wavelengths within
the following ranges at NIST
reference conditions:
1
Size (H × W × D)
Weight
12.5 cm × 3.6 cm × 28.2 cm
0.63 kg
23 °C ± 5 °C
(4 15/16 in × 1 7/16 in × 11
(1.4 lb)
(73 °F ± 9 °F)
/
8
in)
Temperature
operating
storage
—40 °C to 70 °C
(—40 °F to 158 °F)
Relative humidity
0 % to 80 % non-condensing
786 nm ± 0.5 nm
(840 to 860) nm ± 0.25 nm
(1280 to 1330) nm ± 0.25 nm
(1520 to 1625) nm ± 0.25 nm
• 9/125 to 62.5/125 µm (B to D) fiber.
• FC, ST or SC connector.
• Fiber-optic adapter (FOA) used
at NIST.
O R D E R IN G IN F O R M ATIO N
IQ S -1 5 0 X -X X -X -X X
Model
• Source spectral width (FWHM)
< 10 nm.
IQS-1502-Q0-B = 5 mm TEC-Ge detector, 2 % uncertainty at (1310 ± 0.1) nm and (1550 ± 0.1) nm, Singlemode 9/125 µm7
IQS-1502-Q1-B = 5 mm TEC-Ge detector, 0.9 % uncertainty calibrated at NIST, Singlemode 9/125 µm
IQS-1502-Q1-C = 5 mm TEC-Ge detector, 0.9 % uncertainty calibrated at NIST, Multimode 50/125 µm
IQS-1502-Q1-D = 5 mm TEC-Ge detector, 0.9 % uncertainty calibrated at NIST, Multimode 62.5/125 µm
b) ± 3 % uncertainty from 750 nm to
1700 nm, 62.5/125 µm (D) fiber,
source spectral width (FWHM)
< 12 nm, and FC/UPC connector
(ceramic ferrule) with FOA-322.
5. See Ordering Information for details.
6. All uncertainties are valid on the
day of calibration after a warm-up
time of 20 minutes, and specified
with a 95 % confidence level.
7. Test jumper FC/UPC only
8. Singlemode only
Test jumper connector
50 = FC/PC9
54 = SC/PC9
58 = FC/APC narrow key8
74 = ST/PC9
Example: IQS-1502-Q1-B-89
88 = SC/APC8
89 = FC/UPC narrow key8
90 = ST/UPC8
91 = SC/UPC8
9. Multimode only
Corporate Headquarters > 400 Godin Avenue, Vanier (Quebec) G1M 2K2 CANADA Tel.: 1 418 683-0211 Fax: 1 418 683-2170 info@exfo.com
EXFO America
EXFO Europe
EXFO Asia-Pacific
EXFO China
4275 Kellway Circle, Suite 122
Addison, TX 75001 USA
Tel.: 1 800 663-3936
Tel.: +33.1.40.83.85.85
Tel.: +65 6333 8241
Tel.: +86 (10) 6849 2738
Fax: 1 972 836-0164
Fax: +33.1.40.83.04.42
Fax: +65 6333 8242
Fax: +86 (10) 6849 2662
Le Dynasteur, 10/12 rue Andras Beck
92366 Meudon la Forêt Cedex FRANCE
SINGAPORE 169876
151 Chin Swee Road, # 03-29 Manhattan House
Beijing New Century Hotel Office Tower, Room 1754-1755
No. 6 Southern Capital Gym Road
Beijing 100044 P. R. CHINA
EXFO is certified ISO 9001 and attests to the quality of these products. This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. EXFO
has made every effort to ensure that the information contained in this specification sheet is accurate. However, we accept no responsibility for any errors or omissions, and we
reserve the right to modify design, characteristics and products at any time without obligation. Units of measurement in this document conform to SI standards and practices.
Contact EXFO for prices and availability or to obtain the phone number of your local EXFO distributor.
For the most recent version of this spec sheet, please go to the EXFO website at http://www.exfo.com/specs
In case of discrepancy, the Web version takes precedence over any printed literature.
SPIQS1500.2AN
©