EXFO Photonic Solutions Div Trash Compactor FTB 5230 User Manual

O P TIC A L S P E C TR U M A N A LYS E R  
FTB-5230  
N E TW O R K TE S TIN G O P TIC AL  
D e s ig n e d fo r firs t-c la s s c o s t-e ffe c tive n e s s  
Characterization of down to 50 GHz DWDM systems  
Built-in functionalities for CWDM testing  
P la tfo rm C o m p a tib ility  
F TB -4 0 0 U n ive rs a l Te s t S ys te m  
Key feature: wavelength and power drift measurements  
Wide wavelength range of 1250 nm to 1650 nm  
Versatile and powerful  
Telecom Test and Measurement  
F TB -5 2 3 0  
Optical Spectrum Analyzer  
Th e Ch o ic e fo r CWD M O S A Te s tin g  
Since CWDM systems generally use uncooled lasers,  
whose typical wavelength drift is 0.1 nm per degree  
Celsius, it is critical to track the evolution and drift of  
the lasers’ central wavelength. And since path  
attenuation can significantly vary within a single  
channel, passband drift also creates power drift. This is  
why it becomes extremely important to monitor—  
simultaneously for all CWDM channelsthe evolution  
of peak power and wavelength over time. The FTB-5230  
Optical Spectrum Analyzer’s Drift mode functionalities  
makes this an easy task.  
Monitoring wavelength and power drift is essential in CWDM systems.  
TO TAL IN -C H AN N E L P O W E R  
M E AS U R E M E N T  
With channels as wide as CWDM channels,  
monitoring peak power is not enough. The chosen  
OSA must be able to track the total in-channel power—  
a feature only offered by EXFOs FTB-5230.  
R U G G E D N E S S AN D O P TIC AL  
P E R F O R MAN C E  
Measuring peak power, integrated power as well as total in-channel power enables an extremely tight  
control of the network.  
The FTB-5230 Optical Spectrum Analyzer is truly  
optimized for DWDM/CWDM applications, and ready  
for the field-testing challenges they bring. Like all of  
EXFOs OSAs, it comes with high-end optical  
specifications, as well as the added and unique  
advantage of being rugged, portable and battery-  
operated.  
TH E TO O LB O X ADVAN TAG E  
EXFOs exclusive ToolBox software suite runs the test  
module applications on the FTB-400 Universal Test  
System. Easy-to-read graphics and clear instructions  
simplify testing and increase productivity in the field.  
All applications supported by the FTB-400 have a  
common graphical user interface.  
O N E -B U TTO N TE S TIN G  
Simplify testing and stay on schedule with one-button  
testing and the FTB-5230s integrated software. If your  
deadline is tomorrow, the last thing you need is  
software that will slow you down. Turn on the unit and  
the last settings are ready for CWDM system  
characterization. Press Start, and within seconds, you  
receive your data.  
Either in the field or at the central office, the FTB-5230 gets the job done.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F TB -5 2 3 0  
Optical Spectrum Analyzer  
a
S P E C IF IC ATIO N S  
NOTES  
Wavelength range (nm)  
1250 to 1650  
0.1  
a. At 23 °C ± 2 °C, after warm-up.  
b. Full width at half maximum.  
c. At 1550 nm.  
b, c  
Resolution bandwidth FWHM  
d, e  
(nm)  
Wavelength uncertainty  
(nm)  
± 0.1  
f
± 0.02  
d. From 1520 nm to 1610 nm.  
e. Typical.  
d
e
Measurement range (dBm)  
10 to –60  
e, g  
Power uncertainty  
(dB)  
d, e  
± 0.4  
f. After user calibration in the same test session,  
within 10 nm from each calibration point.  
Optical rejection ratio  
Optical rejection ratio  
at 100 GHz (dBc)  
at 50 GHz (dBc)  
45  
d, e  
40  
g. At 1550 nm, –10 dBm.  
e
Scanning time (s)  
4 (35 nm span, multiple-peak analysis)  
c, e  
PDL  
(dB)  
± 0.1  
ORL (dB)  
40  
G E N E R AL S P E C IF IC ATIO N S  
Temperature  
operating  
storage  
0 °C to 40 °C  
(32 °F to 104 °F)  
(4 °F to 120 °F)  
20 °C to 50 °C  
Relative humidity  
Connectors  
0 % to 95 % non-condensing  
EI (EXFO UPC Universal Interface)  
EA (EXFO APC Universal Interface)  
960 mm x 760 mm x 260 mm  
2 kg  
Size (H x W x D) (module)  
Weight (module)  
(3 3/4 in x 3 in x 10 1/4 in)  
(4 lb)  
O R D E R IN G IN F O R M ATIO N  
F TB -5 2 3 0 -X X  
Connector  
EI-EUI-28 = UPC/DIN 47256  
EI-EUI-76 = UPC/HMS-10/AG  
EI-EUI-89 = UPC/FC narrow key  
EI-EUI-90 = UPC/ST  
EA-EUI-28 = APC/DIN 47256  
EA-EUI-89 = APC/FC narrow key  
EA-EUI-91 = APC/SC  
EA-EUI-95 = APC/E-2000  
EI-EUI-91 = UPC/SC  
Example: FTB-5230-EI-EUI-89  
EI-EUI-95 = UPC/E-2000  
EXFO is certified ISO 9001 and attests to the quality of these products.  
This device complies with Part 15 of the FCC Rules. Operation is  
subject to the following two conditions: (1) this device may not cause  
harmful interference, and (2) this device must accept any interference  
received, including interference that may cause undesired operation.  
EXFO has made every effort to ensure that the information contained  
in this specification sheet is accurate. All of EXFO’s manufactured  
products are compliant with the European Unions WEEE directive. For  
more information, please visit www.EXFO.com/recycle. However, we  
accept no responsibility for any errors or omissions, and we reserve the  
right to modify design, characteristics and products at any time without  
obligation. Units of measurement in this document conform to SI  
standards and practices. Contact EXFO for prices and availability or to  
obtain the phone number of your local EXFO distributor.  
Find out more about EXFO's extensive line of high-performance portable instruments by visiting our website at www.EXFO.com.  
EXFO Corporate Headquarters > 400 Godin Avenue, Quebec City (Quebec) G1M 2K2 CANADA Tel.: 1 418 683-0211 Fax: 1 418 683-2170 info@EXFO.com  
Toll-free: 1 800 663-3936 (USA and Canada) www.EXFO.com  
EXFO Montreal  
EXFO Toronto  
EXFO America  
EXFO Europe  
EXFO Asia  
2650 Marie-Curie  
St-Laurent (Quebec) H4S 2C3 CANADA  
Concord (Ontario) L4K 3E5 CANADA  
Plano, TX 75075 USA  
Tel.: 1 514 856-2222  
Tel.: 1 905 738-3741  
Tel.: 1 800 663-3936  
Tel.: +44 2380 246810  
Tel.: +65 6333 8241  
Tel.: +86 (755) 8203 2300  
Fax: 1 514 856-2232  
Fax: 1 905 738-3712  
Fax: 1 972 836-0164  
Fax: +44 2380 246801  
Fax: +65 6333 8242  
Fax: +86 (755) 8203 2306  
160 Drumlin Circle  
3701 Plano Parkway, Suite 160  
For the most recent version of this spec sheet, please go to the EXFO  
website at http://www.EXFO.com/specs  
SOUTHAMPTON > Omega Enterprise Park, Electron Way  
151 Chin Swee Road, #03-29 Manhattan House  
No.88 Fuhua, First Road  
Chandlers Ford, Hampshire S053 4SE ENGLAND  
SINGAPORE 169876  
In case of discrepancy, the Web version takes precedence over any printed  
literature.  
EXFO China  
Shenzhen 518048, CHINA  
Central Tower, Room 801, Futian District  
Beijing New Century Hotel Office Tower, Room 1754-1755  
No. 6 Southern Capital Gym Road  
Beijing 100044 P. R. CHINA  
Tel.: +86 (10) 6849 2738  
Fax: +86 (10) 6849 2662  
SPFTB5230.2AN  
©
2006 EXFO Electro-Optical Engineering Inc. All rights reserved.  
Printed in Canada 06/ 07  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Clarion CD Player DRZ 9255 User Manual
ConnectPRO Switch AVS 12 I User Manual
Craftsman Snow Blower 247884551 User Manual
Curtis Clock Radio RCD951 User Manual
Danby Refrigerator DFF100A2WDB User Manual
Dimplex Indoor Fireplace DFI2309 User Manual
Duracell Battery Charger Ultra CR V3 User Manual
Electro Voice Portable Speaker FR200B User Manual
FieldServer Computer Drive FS 8700 17 User Manual
Fisher Price Video Games V8311 User Manual