Compaq Network Router MTEK6000 User Manual

MTEK6000 SERIES USER'S MANUAL  
MTEK6000 SERIES  
Electronic Flow Corrector & Monitoring  
Devices  
Installation and Operating Instructions  
January 2002  
Part # 900315  
January 2002  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
Analog Sampling ……..................................................................……..…. 3-8  
Special Key Combinations …………………............................................... 3-9  
Assigning The Number of Displayed Digits .............................................. 3-9  
Viewing and Clearing Alarms from the Keypad ....................................... 3-9  
Calibration Mode ........................................................................................... 3-10  
Calibrating the Pressure Transducer ....................................................... 3-10  
Calibrating the Temperature Transducer .................................................. 3-11  
Calibrating the Differential Pressure Transducer ...................................... 3-12  
Section 4: Optional Equipment  
4-1  
Analog Output …………...............................................................................  
4-1  
MTEK6000 Analog Output Specifications …...................................................... 4-3  
Installing the Analog Output Loop ….............................................................  
Calibrating the Analog Output …................................…...............................  
4-3  
4-3  
Section 5: Maintenance and Software Packages  
5-1  
Enclosure Maintenance ...............................................................................  
Changing the Battery ...................................................................................  
5-1  
5-1  
Calibration .................................................................................................... 5-1  
PcGas Meter Reader …................................................................................. 5-2  
PcGas Meter Utility ……................................................................................  
5-2  
PcGas Customer Monitor .............................................................................. 5-2  
MTEK Manager.................................................................................... ......... 5-2  
DC2000 ........................................................................................................  
5-3  
Appendix A : Process Configuration Standard  
Appendix B : Calculations  
Appendix C : Parameter Description  
Appendix D : Board Jumper Positions  
Appendix E : Certifications (CSA, UL, FCC drawings / statements)  
Appendix F : Warranty Information  
A-1  
B-1  
C-1  
D-1  
E-1  
F-1  
G-1  
H-1  
Appendix G : TII Telephone Circuit Surge Suppressor  
Appendix H : Hazardous Area Installation Control Drawings  
January 2002  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
List of Figures  
Figure 1-1, MTEK6000 Exterior View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Figure 1-2, MTEK6000 Interior View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Figure 2-1, Power Connection and Configuration . . . . . . . . . . . . . . . . . . . . . . .  
Figure 2-2, Index Box Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Figure 2-3, Base Plate showing unit and index rotation . . . . . . . . . . . . . . . . . .  
Figure 2-4, Uncorrected counter masking for 8-digit counter . . . . . . . . . . . . . .  
Figure 2-5, Wall Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Figure 2-6, Pipe Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Figure 2-7, Pressure Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  
1-3  
1-3  
2-1  
2-3  
2-4  
2-4  
2-6  
2-6  
2-7  
Figure 2-8, Typical Installation for MTEK6000 EFC . . . . . . . . . . . . . . . . . . . . . . 2-8  
Figure 2-9, Typical Installation for MTEK6000 EFM . . . . . . . . . . . . . . . . . . . . . . 2-8  
Figure 2-10, Temperature Probe Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9  
Figure 2-11, Pulse Output Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10  
Figure 2-12, Corrector Board Connector and Jumper Configuration Diagram . . 2-14  
Figure 3-1, Optional Keypad and Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7  
Figure 4-1, Analog Output Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2  
List of Tables  
Table 2-1, Thermowell Part Numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9  
Table 2-2, Activity Indicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13  
Table A-1, Display mode and function keys for MTEK6000 EFC . . . . . . . . . . .  
A-1  
Table A-2, Standard alarms for MTEK6000 EFC . . . . . . . . . . . . . . . . . . . . . . . . A-1  
Table A-3, Historical data stored in MTEK6000 . . . . . . . . . . . . . . . . . . . . . . . . . A-1  
Table A-4, Display mode and function keys with auxillary pressure . . . . . . . . .  
Table A-5, Standard alarms for MTEK6000 EFC w/ aux. pressure . . . . . . . . . .  
Table A-6, Historical data stored in MTEK6000 EFC w/ aux. pressure . . . . . . .  
A-2  
A-2  
A-2  
Table A-7, Display mode and func. keys with 2 auxillary pressure . . . . . . . . . . . A-3  
Table A-8, Standard alarms for MTEK6000 EFCV w/ 2 aux. pressure . . . . . . . . A-3  
Table A-10, Standard display mode & func. keys for MTEK6000 EFC2  
with 2 auxillary pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-4  
Table A-11, Standard alarms for MTEK6000 EFC2 w/ 2 aux. pressure . . . . . . . A-5  
Tabel A-12, Historical data stored in MTEK6000 EFC2 with 2 aux. pressure . . . A-5  
Table A-13, Display mode & func. keys for MTEK6000 EFM w/ aux. pressure . . A-6  
Table A-14, Standard alarms for MTEK6000 EFM w/ aux. pressure . . . . . . . . . . A-6  
Table A-15, Historical data stored in MTEK6000 EFM w/ aux. pressure . . . . . . . A-6  
Table A-16, Display mode & func. keys for MTEK6000 EPR . . . . . . . . . . . . . . . . A-6  
Table A-17, Standard alarms for MTEK6000 EPR . . . . . . . . . . . . . . . . . . . . . . . . A-7  
Table A-18, Historical data stored in MTEK6000 EPR . . . . . . . . . . . . . . . . . . . . . A-7  
Table A-19, Display mode & func. keys for MTEK6000 ETR . . . . . . . . . . . . . . . . A-8  
Table A-20, Standard alarms for MTEK6000 ETR . . . . . . . . . . . . . . . . . . . . . . . . A-8  
Table A-21, Historical data stored in MTEK6000 ETR . . . . . . . . . . . . . . . . . . . . . A-8  
Table A-22, Display mode & func. keys for MTEK6000 EPTR . . . . . . . . . . . . . . . A-9  
Table A-23, Standard alarms for MTEK6000 EPTR . . . . . . . . . . . . . . . . . . . . . . . A-9  
Table A-24, Historical data stored in MTEK6000 EPTR . . . . . . . . . . . . . . . . . . . . A-9  
January 2002  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
Metretek, Inc. is a registered trademark and MTEK6000, MTEKManager, pcGas, Meter  
Reader, Customer Monitor, AutoPoll, Label Changer, Site I.D. Changer, Units Changer  
and Virtual Keypad are trademarks of Metretek, Inc. All other trademarks are the  
property of organizations not connected with Metretek, Inc. and are used for reference  
purposes only. All contents and specifications in this manual are subject to change  
without notice.  
January 2002  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
January 2002  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
The MTEK6000 product line consists of six  
models that cover a wide range of  
applications.  
CHAPTER 1: OVERVIEW  
NOTE: The MTEK6000 is similar in  
many respects to the Metretek AE6000  
but there are also differences.  
information in this manual applies only to  
the MTEK6000.  
The  
The MTEK6000 EFCV provides a live  
pressure and temperature reading in  
volume corrections for Positive  
Displacement meters (turbine, etc.)  
The MTEK6000 series products are low-  
cost microprocessor-controlled, electronic  
devices for measuring gas flow and volumes  
or monitoring pressure and temperature for a  
The MTEK6000 EFCP provides a live  
pressure and a fixed temperature reading  
in volume corrections for Positive  
Displacement meters (turbine, etc.)  
system.  
With integral pressure and  
temperature transducers, the MTEK6000 is  
designed for accuracy, reliability, and ease  
of maintenance. It can mount directly on a  
meter's index plate, on a wall or pipe.  
The MTEK6000 EFM provides a live  
pressure and temperature reading in  
volume corrections for Orifice meters.  
Low-power CMOS design and sophisticated  
power conservation circuitry allow the  
MTEK6000 to operate one to two years on  
battery power.  
The MTEK6000 EPR provides a live  
pressure reading for monitoring systems.  
The MTEK6000 ETR provides a live  
temperature reading for monitoring  
systems.  
Two pulse inputs, two status inputs, two  
pulse outputs, two external analog inputs  
(4-20 mA or 1-5V) and a tamper input are  
standard. A large 13-digit LCD display,  
with a magnetic scroll switch located on the  
enclosure exterior, permits data viewing  
without opening the enclosure. Station  
parameter display and alarm display can be  
performed with only external device. The  
standard internal 2400/1200/300 baud  
modem provides remote configuration,  
calibration, and retrieval of data.  
The MTEK6000 EPTR provides a live  
pressure and temperature reading for  
monitoring systems.  
HAZARDOUS LOCATIONS  
The MTEK6000 is listed by Underwriter’s  
Laboratories to bear the UL/C-UL mark (US  
& Canadian Listing) for use in hazardous  
locations.  
The operator can also use an industry-  
standard portable computer running  
Metretek software to configure parameters.  
See Chapter 5 for information on these  
The intrinsically safe version can be  
installed in a Class I, Division 1, Group D  
hazardous location when connected through  
the intrinsic safety barriers as listed in  
control drawing 401061. A copy of this  
drawing is shown in appendix E.  
packages.  
The MTEK6000 calculates  
corrected volume using AGA-7, AGA-3,  
AGA-5 and NX-9 or AGA-8 reports.  
The non-incendive version can be installed  
in a Class I, Division II, Group D hazardous  
location when installed in accordance with  
control drawing 401060.  
Optional equipment includes an external  
keypad and display for configuration and  
calibration, analog output modules, two  
additional pulse outputs, serial port modules,  
and various security options.  
This drawing is also listed in Appendix E.  
January 2002  
1 - 1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
WARNING  
Substitution of components may impair  
suitability for Class 1, Division 1 and Class  
1 Division 2 applications.  
COMPLIANCE  
The MTEK6000 device complies with Part  
15 and Part 68 of the FCC Rule.  
See Appendix E for details.  
ONE-YEAR WARRANTY  
Metretek, Inc. warrants the products it  
manufactures to be totally free from any  
defects in materials and workmanship under  
normal operation and use. Metretek, Inc.  
agrees to repair or replace any instrument,  
which is defective due to faulty  
workmanship or material if returned to our  
factory with shipping charges prepaid,  
within one year of original purchase. See  
Appendix F in the back of the manual for  
full warranty details.  
SECURITY OPTIONS  
The MTEK6000 comes standard with wire  
seal screws for the enclosure. Various  
security options are available to prevent  
unauthorized entry into the enclosure:  
A. Tamper screws  
B. Padlock quick release latch  
C. Padlock quick release latch and door ajar  
switch  
D. Door ajar (tamper) alarm  
1 - 2  
January 2002  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
Figure 1-1 MTEK6000 exterior view  
January 2002  
1 - 3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
Figure 1-2 MTEK6000 interior view  
1 - 4  
January 2002  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
5. If the unit is battery powered, verify that  
jumpers JP 19, 20 and 24 are in the A to B  
position. Load the battery pack with fresh  
batteries and connect it to the corrector at  
position J6. Repeat with the second battery  
pack connecting it at position J7 (see Fig.  
2-1). Go to step 7.  
Chapter 2: Installation  
UNPACKING  
1. Thoroughly examine the box to verify it  
was not damaged in shipping. If you find  
damage, immediately file a claim with the  
shipper.  
2. Carefully unpack the MTEK6000 from  
the shipping container. Verify that the  
box contains every item listed on the  
shipping order.  
INITIAL CHECK OUT  
!!! CAUTION !!!  
This unit contains certain electronic  
components that are sensitive to  
electrostatic discharge (ESD); therefore,  
proper precautions should be taken  
during maintenance operations to avoid  
Figure 2- 1 Power connection and configuration  
6. If the unit is powered via an external  
power source (UPS or SPS option), make  
sure that jumpers JP 19, 20 and 24 are in  
the B to C position and connect the DC  
power input to the DC input screw  
terminals 6 (V+) and 5 (GND) prior to  
power up. Refer to Fig. 2-1 for locating  
the connection points  
ESD.  
It is recommended that the  
operator first touch the shell of the MS  
connector (RS-232C port) on the left side  
of the unit to dissipate any accumulated  
static charge. Additional precautions  
may be taken in order to minimize the  
possibility of ESD, including the use of a  
grounding wrist strap (i.e., 3M part  
number 2214).  
WARNING  
The operating voltage range from an external  
power source is 7-18 VDC. Do NOT exceed  
recommended input voltage of 18 VDC.  
3. Examine the label on the left side of the  
enclosure. It indicates the serial number  
and pressure range for your unit. Verify  
that these parameters match your  
requirements. If they do not, please  
contact your sales representative.  
7. When you first apply power, the display  
will show the first two capital letters of the  
first label, followed by the value and then  
the units.  
For example, the label  
Corrected Volume MCF with a value of  
00000000 would be displayed as CV  
00000000 MCF.  
4. Open the front door by loosening the  
upper and lower right hand corner screws  
or quick release latches of the enclosure  
and swinging the door out. See Fig. 1-1.  
January 2002  
2-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
B. MTEK6000 UPS power supply - an un-  
interruptible 12 VDC power supply with  
battery back up.  
NOTE  
The flashing LCD display indicates an  
alarm condition (e.g. First Time Power).  
See Chapter 3 for information on alarms.  
C. SPS 50 solar system - 10 to 64 W systems  
available with battery backup; while  
selected system size depends on  
geographic location, degree of sun  
exposure, equipment power consumption,  
and site obstructions, most MTEK6000  
applications only require a 10W system.  
8. You can now view selected parameter  
values on the display by using the scroll  
switch. The scroll switch is activated by  
the use of a magnet (one is shipped with  
the unit). See Display Mode, in Chapter  
3, for information about this function.  
POWER FOR THE MTEK6000  
MOUNTING THE INSTRUMENT ON  
THE METER  
Two 4.5V alkaline battery packs (part #  
1011-0035C-001) supply operating power to  
the device for approximately two years of  
typical operation. Recommended operating  
temperature range for the MTEK6000 when  
powered with these packs is –4 F (-20 C) to  
130 F (54 C). If this supply should fail, an  
on-board back-up battery will maintain the  
unit's memory and real time clock. Backup  
power can maintain history data for up to  
seven years. When back-up power is used,  
the unit discontinues normal operation until  
the main battery pack is replaced.  
1. Check the meter's rotation direction.  
Standard setup is clockwise rotation of the  
meter output shaft, as viewed from the top.  
The rotation of the unit can be changed to  
counterclockwise. Also, the input drive  
value for the unit can be changed.  
2. Align the instrument's index base plate  
holes with the corresponding holes in the  
meter's index base plate. Secure the unit  
by bolting it to the meter. Ensure that the  
drive dogs are correctly aligned and not  
binding.  
Note that only one power source powers  
the MTEK6000; connection of battery  
packs to a unit configured for external  
power does not provide an additional  
source of backup power for the unit.  
3. Plug all unused holes in the index base  
plate with the caps provided in the  
accessory package.  
Optional Power Supplies  
Several optional supplies are available:  
A. Two Single-D Lithium battery packs -  
(part # 1011-0039B-001) provides an  
approximate life expectancy of two years  
over the temperature range -220 F (-300 C)  
0
to 1580 F (70 C). Each 3.6V, 13.0 AH  
battery pack can be used individually (~  
1yr life) or as a pair.  
2-2  
January 2002  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
SETTING UP THE INDEX ASSEMBLY  
Output Shaft Rotation  
feet/revolution (ft3/rev) and 0.1, 1 and 10  
cubic meters/revolution (m3/rev). To change  
the input drive value to 5 ft3/rev:  
To change the rotation of the output shaft to  
counterclockwise (figure 2-2):  
1. Loosen set screw a on compound gear A.  
1. Loosen set screw e on gear E.  
2. Disengage gear E from counter gear D.  
3. Tighten set screw e.  
2. Lower gear A until its upper teeth engage  
the upper teeth of compound gear B.  
3. Tighten set screw a.  
4. Take care to align the gears properly, and  
verify that they turn freely and do not  
bind.  
4. Loosen set screw c on gear C.  
5. Engage gear C with counter gear D.  
6. Tighten set crew c.  
When changing a MTEK6000 in the field  
from a 10, 100, 1,000 or 10,000 ft3/rev drive  
meter with a 5 ft3/rev drive, the CF per Pulse  
In or Meter Drive parameter, must be  
changed to a value of 10 using software or the  
optional, external keypad / display.  
7. Take care to align the gears properly, and  
verify that they turn freely and do not  
bind.  
NOTE  
Electronic parameter CF per Pulse In or  
Meter Drive should always equal the meter  
drive value EXCEPT for 5 ft3/rev meters  
which should be set at 10. When switching to  
5 ft3/rev, the gear ratio is adjusted so that two  
revolutions of the input drive gear result in  
one revolution of the magnet and hub  
assembly, which sends a pulse signal to the  
instrument that equals 10 ft3/rev. The value  
can be changed using software or the  
optional, external keypad / display (see  
Appendix A for addressing).  
To change the input drive value to 10, 100,  
1,000 or 10,000 ft3/rev, return compound gear  
A to its original, factory-set position. Change  
the CF Per Pulse In or Meter Drive  
parameter using software or the optional,  
external keypad / display (see Appendix A for  
addressing).  
Figure 2- 2 Index Box Assembly  
Input Drive Value  
Fig. 2-2 above shows gearing in the correct  
position for 10, 100, 1,000 or 10,000 cubic  
January 2002  
2-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
NOTE  
Figure 2- 3 Base Plate Showing Unit &  
Index Rotation  
When you have changed the input drive  
value, be sure to remove the existing drive  
value label from the window and replace it  
UNIT AND INDEX ROTATION  
with a new label that states the current input  
drive value.  
For your convenience,  
In general, the label side of the index base  
plate (front) should face the front of the meter.  
This allows the MTEK6000 EFC to also face  
the front of the meter. In certain applications,  
the MTEK6000 EFC and index can be  
installed 1800 from the standard position so  
that viewing of the MTEK6000 EFC is  
acceptable. To rotate the unit and index, refer  
to Fig. 2-3 and do the following:  
Metretek, Inc. supplies extra labels with the  
MTEK6000 EFC.  
Magnet Sensor Adjustment  
The index box assembly contains a reed switch  
(G) and a corresponding magnet (F). The magnet  
should be positioned so there is 0.07”-0.01”  
clearance between the magnet and switch  
(1.78mm-0.25mm). To adjust, refer to Figure 2-2  
and do the following:  
1. Remove the 4 bolts at location A.  
2. Rotate the base plate 1800 clockwise so  
that the front label side is now facing the  
rear of the unit.  
1. Loosen set screw h.  
2. Move the magnet until clearance is  
correct.  
3. Insert the 4 bolts into location B with the  
two lockwashers as shown. Fig. 2-3  
shows the index orientation after the  
3. Retighten set screw h.  
4. Verify the clearance is still correct.  
If not, repeat the procedure.  
rotation is completed.  
OVERTIGHTEN).  
(DO NOT  
COUNTER MASKING  
Figure 2- 4 Uncorrected Counter  
Masking for Eight Digit Counter  
2-4  
January 2002  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
PULSE INPUT TO THE MTEK6000  
Software is used to configure the pulse input.  
The optional, external keypad / display can  
also be used to configure the pulse input and  
other parameters. See Appendix A for  
addressing.  
Magnetically operated reed switches inside  
the meter drive assembly send electronic  
pulses as the drive turns. These pulses  
represent uncorrected meter volume to the  
instrument.  
WALL AND PIPE MOUNTING  
To eliminate false counts that can result from  
the reed switch "bounce”, the MTEK6000  
The MTEK6000 can also mount directly on a  
wall or on a pipe. Mounting feet are provided  
for wall mounting. See Figure 2-5.  
uses  
a
set/reset,  
dual-reed  
switch  
configuration. An input pulse is generated  
only when the opening and closing of the  
first switch is followed by the opening and  
closing of the second switch. The main  
counter input is also monitored for fault  
conditions. When enabled, if any of the dual-  
reed switches should be defective, the input  
pulses will automatically switch to the  
working counter input and the MTEK6000  
will generate a Faulty Counter alarm. This  
function is enabled if the Counter Fault  
Monitoring parameter is set to 1 and  
disabled if set to 0. The default value is 0 for  
disable. See Appendix A for addressing.  
For pipe mounting, approximately 10 feet of  
2-inch rigid iron pipe or conduit is required.  
The pipe should be installed 18 to 24 inches in  
the ground in 6 inches wide concrete. The  
length of the 2-inch mounting pipe or conduit  
will vary according to the site, but typical  
installations place the MTEK6000 at about  
eye level for ease of operation. Mounting  
plates are provided for pipe mounting. Secure  
the MTEK6000 to the pipe with the provided  
U-bolts, washers and hex nuts. See  
Figure 2-6.  
For units supplied with indexes, the main  
counter input is terminated at the UNIT  
(BLK) MTA connector and the uncorrected  
pulse wiring at the FIELD (WHT) MTA  
connector at the lower right hand corner of  
the corrector printed circuit assembly.  
Metretek, Inc. can supply a remote index  
similar to the main index or a sandwich  
pulser for wall or pipe mount installations.  
When the sandwich or external pulser option  
(1, 10, 20, or 50 pulse per revolution) is  
supplied, connect the N.O., COM, and N.C.  
wiring to terminals 32 (SET1), 33 (GND) and  
34 (RST1) respectively.  
The pulse input is software selectable for  
Form C (three-wire) or Form A (two-wire)  
connection.  
January 2002  
2-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
Figure 2- 5 Wall Mounting  
Figure 2- 4 Pipe Mounting  
2-6  
January 2002  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
TRANSDUCERS IN THE MTEK6000  
The MTEK6000 uses a precision strain gauge  
pressure transducer mounted inside the unit,  
combining maximum accuracy with low  
power consumption.  
To sense gas  
temperature, the MTEK6000 employs a  
highly linear and stable device, a platinum  
resistive temperature detector (RTD). Case  
temperature sensing is accomplished with an  
on-board precision reference integrated  
circuit (IC).  
CONNECTING THE PRESSURE  
TUBING  
Figure 2- 5 Pressure Connections  
WARNING  
You must DEPRESSURIZE THE METER  
and its associated piping before you make  
pressure piping connections. FAILURE to  
do so may result in EXPLOSION and FIRE,  
causing SERIOUS PERSONAL INJURY  
and PROPERTY DAMAGE.  
As a minimum requirement, connect the  
pressure tubing as shown in Fig. 2-7. An  
optional pressure tubing connection kit (part #  
2019-0009B-001) can be shipped with each  
instrument. Fig. 2-8 is the recommended  
pressure installation for ease of operation.  
Additional piping and valve are required for  
the installation and are not supplied. Use  
Teflon tape or pipe seal compound on all  
threaded connections. The tubing supplied in  
the kit may be longer than you need for your  
installation. You can cut or coil the tubing,  
but do not make any sharp bends in it  
(minimum radius is 3/4"). Tighten all the  
connections and perform a leak test once the  
meter and instrument are pressurized.  
Do NOT attempt to connect any piping or  
fittings to a meter or pipe under pressure.  
Do NOT SMOKE while connecting gas or  
test pressure to the meter.  
Note  
Federal Standard 192.02 requires a shut-off  
valve between the pressure source and the  
instrument. A needle valve is supplied with  
the optional pressure piping kit for this  
purpose. Its rating is 1,500 PSIG MAOP.  
January 2002  
2-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
Figure 2- 7 Typical Installation for MTEK6000 EFC  
Figure 2- 6 Typical Installation for MTEK6000 EFM  
2-8  
January 2002  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
INSTALLING THE THERMAL PROBE  
A thermal (temperature) probe is connected  
to the MTEK6000 by a 6-foot (2-meter)  
cable. You should coil excess cable to  
prevent possible damage. The probe is  
designed to fit into standard Metretek, Inc.  
thermowells. Optional 15-foot (4.5 meter)  
and 30-foot (9-meter) cables are available.  
See Fig 2-10.  
To install the thermal probe, use the supplied  
temperature probe adapter. Refer to Table  
2-1. Insert the probe into the thermowell and  
tighten the securing nut FINGER TIGHT  
only. The standard adapter is a ½” fitting.  
Users retrofitting instruments requiring the  
5/8” adapter can order the adapter from  
Metretek, Inc.  
Oil or ethylene glycol (antifreeze) should be  
used to improve heat transfer from the  
thermowell to the thermal probe. Be aware,  
however, that it is possible to cause hydraulic  
crushing of the probe. This can happen when  
there is little or no air in the thermowell  
above the probe. When the probe is fastened  
by tightening the securing nut, the space in  
the well decreases as the probe enters. As a  
result, hydraulic pressure may rise high  
enough to cause damage. If you use oil or  
antifreeze, make sure there is enough air  
in the thermowell above the fluid to  
prevent crushing the probe.  
Figure 2- 8 Temperature Probe Connection  
Table 2-1: Thermowell Part Numbers  
Pipe Size  
Insertion  
Length  
Thermowell  
Part Number  
4 in.  
2 ½ in.  
4 ½ in.  
7 ½ in.  
10 ½ in.  
5340-0373 ½” NPT  
5340-0377 ¾” NPT  
5340-0384 1” NPT  
6 in.  
8 in.  
5340-0372 ½” NPT  
5340-0376 ¾” NPT  
5340-0383 1” NPT  
It is recommended that the thermowell be  
installed in the meter outlet pipe one or two  
diameters from the meter outlet.  
The  
12 in.  
14 in.  
5340-0371 ½” NPT  
5340-0375 ¾” NPT  
5340-0379 1” NPT  
insertion length of the thermowell must be  
sufficient to extend at least to the pipe center,  
but no further than 75% of the pipe’s  
diameter. Thermowells should not be situated  
where they will be exposed to direct sunlight.  
A sunshield should be used for installations  
where this cannot be avoided.  
16 in.  
20 in.  
5340-0370 ½” NPT  
5340-0374 ¾” NPT  
5340-0378 1” NPT  
January 2002  
2-9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
INSTALLING THE PULSE OUTPUT  
WIRING  
configuration requires that a constant wetting  
current be available at both the normally open  
and normally closed terminals by the device  
attached to that pulse output.  
The MTEK6000 comes standard with a  
board installed that provides two optically  
isolated pulse outputs. These outputs are  
configurable as either Form C or Form A  
type outputs. An alternative version of the  
board is available that provides four pulse  
outputs. Both versions of the board also  
provide terminal block positions to access the  
uncorrected mechanical volume switch  
output of the index. See Fig. 2-11 for pulse  
output wiring. The option boards optical  
coupling and physical arrangement of  
circuitry provide a minimum of 1,500 volts  
of isolation.  
The corrected volume pulse output generated  
can be scaled to any desired volume value.  
Typical values are 10, 100, 1,000, or 10,000  
cubic feet per pulse, or the metric equivalents.  
The scaling factor is selected by the Pulse  
Out CF Per Pulse parameters. The pulse  
duration (width) is also configurable up to  
5,000 ms. The Pulse Output On-Time and  
Off-Time parameters determine the pulse  
time for corrected volume, uncorrected  
volume and pressure corrected volume pulses.  
The Alarm Pulse Time (ms) parameter  
determines the pulse time for alarm outputs.  
See Appendix C in this manual or Meter  
Reader Help screen for description of this  
parameter.  
Wiring connections for the pulse outputs are  
made from terminals 19 to 17 for Pulse  
output #1, terminals 16 to 14 for Pulse output  
#2, terminals 13 to 11 for Pulse output #3 and  
terminals 10 to 8 for Pulse output #4. See  
Fig. 2-11 for pulse output wiring location and  
Form C vs. Form A jumpering. Note that  
proper operation of the Form C pulse output  
Figure 2- 9 Pulse Output Wiring  
2-10  
January 2002  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
Volume and Alarm Pulse Specifications  
Uncorrected Pulse Output Specifications  
1. All pulse outputs are isolated from  
ground and each other. Provides 1,500  
volts between input and output and  
between contact sets.  
1. 3W contact rating (power dissipation).  
2. Maximum switching voltage up to 30V.  
3. Maximum switching current up to 200mA.  
4. Maximum continuous current @ 500 mA.  
2. Form C: DC load only, 125mWdc max,  
50Vdc max  
Form A: AC or DC load 800mW max,  
400V max., 100mA max, continuous  
3. Configurable pulse width from 1 to 5,000  
milliseconds (ms).  
NOTE  
All pulse outputs are disabled in the standard  
unit to conserve power. The Alarm pulse  
output is a one time pulse output signal. No  
other alarm pulse will be generated until the  
alarm is cleared and becomes active again.  
Uncorrected Mechanical Pulse Output  
(Units with a Metretek Index)  
The uncorrected mechanical pulse output is  
derived from the Form C reed switch in the  
index assembly. As the magnet in the drive's  
assembly rotates past the Form C switch, a  
single uncorrected volume pulse output is  
generated. Volume per pulse is determined  
by the drive rate. Each uncorrected volume  
pulse is equal to the gas flow for one shaft  
revolution.  
The pulse output can be wired as a Form A or  
Form C switch output. Use terminals 25  
(normally open), 24 (common) and 23  
(normally closed) for Form C output. To  
wire as Form A, use terminals 25 (normally  
open) and 24 (common) and make no  
connection to terminal 23 (normally closed).  
The duration (width) of each pulse is equal to  
the length of time the reed switch is in its  
closed position (depends on the rate of the  
meter). No configuration is necessary to  
enable the uncorrected mechanical pulse  
output.  
January 2002  
2-11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
COMMUNICATIONS  
Modem Communications (2400 Baud)  
To communicate with the MTEK6000, the  
Site ID (RUID) in the device must be the  
same as the Site ID entered in the software  
NOTE  
This modem complies with Part 68 of the  
FCC Rules. See Appendix E for details.  
package.  
The Site ID is a unique  
identification number (1 to 65,535) that  
allows the Metretek, Inc. software packages  
to communicate with the MTEK6000. The  
default Site ID number is 1. Software can be  
used to enter a number other than the default.  
Refer to the respective software User’s  
Manual for additional information on these  
and other functions. The optional external  
display and keypad can also be used to  
change the Site ID from its default value.  
The internal Hayes compatible modem offers  
automatic answering and dialing. The modem  
communicates at 2400/1200/300 baud. The  
modem by itself can only be used in areas  
classified as non-hazardous or Class I,  
Division 2. To maintain the MTEK6000’s  
intrinsic safety classification in more  
hazardous areas such as Class I, Division 1, an  
optional Phone Line Interface (PLI) must be  
used. This device removes the high voltage  
ring-detect circuitry from the device and  
brings low-level signals into the hazardous  
area through intrinsic safety barriers.  
RS-232C Serial (Direct) Communications  
(9600 Baud)  
In its standard configuration, the MTEK6000  
is equipped with one RS-232C serial port.  
An optional RS-232C serial cable (Part #  
1002-0235C-001) is required for direct  
communications. The serial port allows an  
operator to configure and collect data with an  
industry-standard (IBM, Compaq, etc.)  
portable computer (software is required for  
Connecting the Telephone Line  
If the MTEK6000 is situated in a non-  
hazardous or Class I, Division 2 area, connect  
the tip and ring wires from the telephone  
company's terminal box to the TIP and RING  
terminals (1 and 2 respectively).  
this  
function).  
The  
MTEK6000  
If it is installed in a Class I, Division 1, Group  
D area, install the unit per the reference  
communicates at 9600 baud with portable or  
host computers connected directly to the  
serial port. When communicating with the  
MTEK6000, Busy will be displayed on the  
display. Table 2-2 shows the diagnostic  
features of the Activity indicator when the  
cable is connected.  
drawing shown Appendix E.  
Also see  
Appendix D for proper jumper settings.  
Installation of the phone line surge protection  
device provided with the MTEK6000 is  
strongly recommended when the MTEK6000s  
internal modem is connected to a telephone  
line. The device is a separate gas tube type  
phone line surge suppressor and is housed in  
its own enclosure suitable for mounting  
directly to a telephone pole or other structure.  
WARNING  
The MTEK6000 will not go to sleep if the  
RS-232C serial cable is left connected and  
battery life will be affected drastically.  
2-12  
January 2002  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
Table 2-2 Activity Indicator  
MTEK6000 Function (RS232 cable connected)  
RS-232C cable connected  
Set #1 Pulse received  
Activity Indicator  
1 long blink  
1 short blink  
Reset #1 Pulse received  
2 short blinks  
Set #2 Pulse received  
3 short blinks  
Reset #2 Pulse received  
4 short blinks  
RS-232C cable disconnected  
3 long blinks after a few seconds delay  
January 2002  
2-13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
Figure 2- 10 Corrector Board connection and jumper configuration diagram  
2-14  
January 2002  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
GROUNDING  
case it is of no benefit to earth ground the  
MTEK6000’s external lug.  
The information presented here is merely a  
guideline to help customers avoid surge damage  
to the MTEK6000. None of these guidelines are  
to be construed as replacing or superseding  
rules and practices defined by the National  
Electrical Code (NEC), or the Classification of  
Gas Utility Areas for Electrical Installations  
guidelines, as published by the American Gas  
Association (AGA) or other regulatory agency.  
A sound understanding of Federal, State and  
Local laws is fundamental to proper and legal  
installation work.  
When the external ground lug of the  
MTEK6000 is to be used, it should be  
connected to a common ground rod (or "bed"  
of grounding equipment) to which is securely  
tied all other equipment chassis, metal  
cabinets, and intrinsic ground brackets. Solid  
copper ground wire or ground strapping of an  
approved size and type must be used to tie this  
equipment to the rod. If possible, it is far  
preferable that all external equipment be tied  
to a single site ground, that the distance  
between the MTEK6000 and external  
equipment is kept at a minimum (less than 20  
feet is best), and the ground rod be located no  
farther than halfway between the MTEK6000  
and the other equipment.  
The MTEK6000 is configured so that the  
majority of the internal metal components  
within the device are connected (common) to  
the gas pipeline / meter to which the  
instrument is attached. Additionally, a large  
surge bypass MOV device has been provided  
inside the MTEK6000 that provides an  
alternate path (rather than through the  
correctors electronics!) for surge current to  
flow. One side of this device connects to the  
pipe through the index & meter. The other  
side is brought out through the enclosure to  
an external copper grounding lug. The  
separation of these two points allows for the  
existence of cathodic protection voltage  
levels on the pipe (typically about 1 volt  
below the surrounding soil) while still  
providing a path for surge current to safely  
bridge these points, find earth ground, and  
not damage the instrument.  
If separate site grounds must be used, as when  
the MTEK6000 and UPS are separated by a  
distance greater than 20 feet, Metretek can  
provide an optional device, the SPM (Surge  
Protection Module). A pair of SPMs can  
properly protect both devices in this  
circumstance. Control drawings detailing  
proper wiring of SPMs (including hazardous  
areas) are included with the SPM.  
REFERENCES  
National Electrical Code (NFPA):  
Article 250 – Grounding  
Articles 500 & 501- Hazardous  
(Classified) Locations  
If the MTEK6000 is to be connected to a  
telephone line (either on-board or through a  
PLI mounted in a safe area) or connected to a  
UPS (AC power supply), then the external  
ground lug provided on the MTEK6000  
should be connected to earth ground. If the  
unit is not making connection to the phone  
co. lines, power co. lines, or other external  
equipment, then connecting the unit to an  
earth ground simply introduces a path for  
surges that otherwise would not exist. In this  
Article 504 - Intrinsically Safe Systems  
The IAEI Soares' Book on Grounding  
(Available through the NFPA)  
PolyPhaser Corp. Catalog of Lightning /  
EMP & Grounding Solutions Minden,  
Nevada (800) 325-7170  
January 2002  
2-15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
2-16  
January 2002  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
Calling the unit via modem.  
Chapter 3 : Operating Modes  
Waking up on a specified number of  
pulses set by the Wake Up On Pulse  
parameter. See Appendix A for the  
Wake Up On Pulse (event driven)  
parameter address and the Wake Up on  
Pulse (event driven) section later in this  
chapter for more details on this feature.  
Waking up on a limit violation set by  
analog minimum or maximum setpoints.  
This is only valid when analog sampling  
is enabled. See Analog Sampling later  
in this chapter for details.  
The MTEK6000 operates in any of five  
modes:  
Sleep  
Display  
Alarm  
Configuration (requires Virtual keypad,  
Meter Reader 4.10 or later, or pcGas  
Host software, or the optional external  
keypad and display)  
Calibration (requires Virtual keypad,  
Meter Reader 4.10 or later, software or  
the optional external keypad and  
display)  
See the section on Alarm Mode later in this  
chapter for information on viewing and  
clearing alarms.  
SLEEP/WAKE-UP MODE  
Once the unit is awake, it will automatically  
power-down in one minute after the operator  
stops interacting with it.  
In normal operation, the MTEK6000  
maintains a powered-down state (sleep  
mode) to conserve battery life. In this state,  
the display will show the first label and the  
most recent value prior to the next wake-up  
NOTE  
The MTEK6000 will not go to sleep if the  
RS-232C serial cable is connected and  
battery life will be affected drastically.  
interval.  
The unit will be updated  
depending upon the user's programmed  
wake-up interval. The default wake-up  
interval is 10 minutes. See Appendix A for a  
list of common MTEK6000 items including  
this parameter. Shorter intervals result in  
more frequent data; longer intervals provide  
for longer battery life. The scheduled wake-  
ups result in immediate power-down after  
performing calculations.  
DISPLAY MODE  
In this mode, the display normally shows a  
two-character label and a value. You can  
view the next displayed label without having  
to open the front door by touching the scroll  
switch with a magnet.  
In addition to the scheduled wake-ups, the  
MTEK6000 can be brought up from its sleep  
mode by any one of the following:  
In display mode, only limited parameters  
with assigned labels can be viewed. See  
Appendix A for a list of display mode  
parameters for your device.  
Connecting a portable computer to the  
unit's serial communication port.  
Scrolling the external display with a  
magnet.  
Pressing any key on the optional keypad.  
Opening the enclosure door. (If tamper  
switch is installed)  
Change in state of the status inputs.  
January 2002  
3 -1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
alarms using the magnet:  
ALARM MODE  
The MTEK6000 can be configured to  
activate an alarm when certain conditions  
are met or when user defined limits are  
exceeded. You can display active alarm  
messages on the optional external keypad  
and display or alarm codes on the standard  
display. The unit can also automatically call  
a host computer running Metretek, Inc.  
software programs to report alarms.  
1. Apply the magnet to the scroll switch  
until the outer display shows AL  
XXXXXX. (See Appendix A for a list of  
alarms and codes for your device). The  
alarms are described later in this chapter.  
2. To view another alarm, apply and  
remove the magnet briefly. The unit  
should advance to the next alarm code if  
other alarms are active.  
A history log is kept in the device on each  
alarm condition consisting of:  
3. To clear an alarm, hold the magnet on  
the scroll switch for approximately five  
seconds until the display flashes OK,  
indicating the alarm was cleared.  
Current value  
Type of alarm (high, low etc.)  
Setpoint value (alarm limit)  
Time of alarm  
Date of alarm  
Time out of alarm  
Date out of alarm  
4. If more than one alarm is active, the  
display will show the next alarm code.  
Apply the magnet to the scroll switch for  
approximately five seconds and clear the  
alarm.  
Extreme value alarm  
5. The user can now view the standard  
display parameters in Appendix A.  
Appendix A shows the standard alarms and  
codes for your device.  
NOTE  
In addition to the standard alarms, you can  
program the MTEK6000 to monitor and  
report on almost any condition, such as  
meter tampering, liquid level, valve status,  
heater status, etc.  
The ability to clear alarms with a magnet is  
default in the MTEK6000. This function  
can be disabled using Metretek, Inc.’s  
software packages or the optional external  
keypad and display. If disabled, the user  
would not have access to clear alarms, only  
to view parameters.  
NOTE  
Additional hardware equipment and  
configuration could be required for custom  
alarm monitoring.  
Current Day Flow Alarm (EFC & EFM)  
If the current day's total volume should  
exceed the Current Day High Volume Alarm  
Setpoint, a Current Day Flow alarm will be  
initiated. The alarm will remain active until  
the value for the current day volume is reset  
the next day at roll time. The setpoints are  
user configurable with default values of  
100000 and 99990 respectively (see  
Appendix A for parameter addressing for  
your device).  
Viewing and clearing alarms using the  
magnet  
Alarms in the MTEK6000 can be  
recognized by the flashing outer display.  
This is an indication that one or more of the  
standard alarms were initiated (see  
Appendix A for a list of alarms and codes  
for your device). To view and clear the  
3-2  
January 2002  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
For transport or interruptible customers, this  
parameter can be used to alarm when an  
account has exceeded a predetermined daily  
volume allocation.  
High Pressure Alarm  
If the gas pressure should exceed the High  
Pressure Alarm Setpoint, a High Pressure  
alarm will be initiated. The alarm will  
remain active until the pressure falls below  
the High Pressure Reset parameter value.  
The setpoints are user configurable with  
default values of 1500 and 1480 respectively  
(see Appendix A for parameter addressing  
for your device).  
Faulty Counter Alarm (EFC)  
This alarm is only valid for EFC units with a  
Form C switch. If any of the dual-reed  
switches in the index assembly fail, pulses  
to the unit would automatically switch to a  
working counter input. When this happens,  
the unit generates a Faulty Counter alarm  
(see the section on Pulse Input to the  
MTEK6000 in chapter 2 for more details).  
High Temperature Alarm  
If the gas flow temperature should exceed  
the High Temperature Alarm Setpoint, a  
High Temperature alarm will be initiated.  
The alarm will remain active until the  
temperature falls below the High  
Temperature Reset parameter value. The  
setpoints are user configurable with default  
values of 200 and 180 respectively (see  
Appendix A for parameter addressing for  
your device).  
First Time Power Alarm  
First Time Power alarm is defined as the re-  
application of power after interruption of the  
power source. For example, whenever the  
battery  
is  
disconnected  
and  
then  
subsequently reconnected, the unit records  
the First Time Power event.  
Low Differential Pressure Alarm (EFM)  
If the differential pressure should fall below  
the Low Differential Pressure Alarm  
Setpoint, a Low Differential Pressure  
alarm will be initiated. The alarm will  
remain active until the differential pressure  
rises above the Low Differential Pressure  
Reset parameter value. The setpoints are  
user configurable with default values of -  
100 and -80 respectively (see Appendix A  
for parameter addressing for your device).  
High Flow Rate Alarm (EFC & EFM)  
If the flow rate should exceed the High Flow  
Rate Alarm Setpoint, a High Flow Rate  
alarm will be initiated. The alarm will  
remain active until the flow rate falls below  
the High Flow Rate Reset parameter value.  
The setpoints are user configurable with  
default values of 100000 and 99990  
respectively (see Appendix A for parameter  
addressing for your device).  
High Differential Pressure Alarm (EFM)  
If the differential pressure should exceed the  
High Differential Pressure Alarm Setpoint, a  
High Differential Pressure alarm will be  
initiated. The alarm will remain active until  
the differential pressure falls below the High  
Differential Pressure Reset parameter value.  
The setpoints are user configurable with  
default values of 1500 and 1480 respectively  
(see Appendix A for parameter addressing  
for your device).  
Low Flow Rate Alarm (EFC & EFM)  
If the flow rate should fall below the Low  
Flow Rate Alarm Setpoint, a Low Flow  
Rate alarm will be initiated. The alarm will  
remain active until the flow rate rises above  
the Low Flow Rate Reset parameter value.  
The setpoints are user configurable with  
default values of -100 and -80 respectively  
(see Appendix A for parameter addressing  
for your device).  
January 2002  
3 -3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
Low Pressure Alarm  
Lost Pressure Alarm  
If the gas pressure should fall below the  
Low Pressure Alarm Setpoint, a Low  
Pressure alarm will be initiated. The alarm  
will remain active until the pressure rises  
above the Low Pressure Reset parameter  
value. The setpoints are user configurable  
with default values of -100 and -80  
respectively (see Appendix A for parameter  
addressing for your device).  
A Lost Pressure alarm is generated when  
the pressure circuitry is over-ranged. This  
can occur if the pressure transducer is  
defective, or disconnected from the analog  
board causing pressure readings to be above  
or below the range of the transducer.  
Lost Temperature Alarm  
A Lost Temperature alarm is generated  
when the temperature circuitry is over-  
ranged. This can occur if the thermal  
(temperature) probe is defective, or  
disconnected from the analog board causing  
temperature readings to be above or below  
the range of the probe.  
Low Supply Volts Alarms  
If the supply voltage to the unit falls below  
the Low Supply Volts Alarm Setpoint value,  
a Low Supply Volts alarm will be initiated.  
The alarm will remain active until the  
supply voltage is greater than the Low  
Supply Volts Alarm Reset parameter. The  
setpoints are user configurable with default  
values of 2.9 and 3.1 volts respectively for  
battery powered units (see Appendix A for  
parameter addressing for your device).  
LowVolt Shutdown  
On battery powered units, if the battery  
voltage decreases to approximately 2.8  
volts, an interrupt will be triggered and the  
unit will enter Low Voltage (Critical)  
Shutdown mode.  
displayed on the display.  
Low batt will be  
This is an  
Low Temperature Alarm  
If the gas flow temperature should fall  
below the Low Temperature Alarm Setpoint,  
a Low Temperature alarm will be initiated.  
The alarm will remain active until the  
temperature rises above the Low  
Temperature Reset parameter value. The  
setpoints are user configurable with default  
values of -100 and -80 respectively (see  
Appendix A for parameter addressing for  
your device).  
indication that the supply voltage is  
absolutely too low to operate the unit  
properly. The battery must be changed at  
this point or adequate supply voltage  
applied.  
For units that are externally  
powered, this point will occur when the  
input voltage drops to approximately 6.0  
volts. In this mode, all operations cease, and  
the unit operates in a protective mode. The  
on-board battery will continue to protect the  
unit's memory; therefore data prior to  
entering this mode will be maintained. The  
supply voltage is monitored constantly and  
the unit will reset itself should the supply  
voltage become greater than 6.0 volts. If the  
unit is left alone without applying adequate  
supply voltage, the battery will continue to  
drain and the outside display will eventually  
go blank.  
Lost Differential Pressure Alarm (EFM)  
A Lost Differential Pressure alarm is  
generated when the differential pressure  
circuitry is over-ranged. This can occur if  
the differential pressure transmitter is  
defective, or disconnected from the analog  
board causing differential pressure readings  
to be above or below the range of the  
transducer.  
When sufficient supply voltage is applied  
and the unit powers-up, a LowVolt  
Shutdown alarm will be recorded.  
3-4  
January 2002  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
# of Record =  
81,000  
.
4 x (# of items to log +1)  
Open Door Alarm (optional)  
NOTE  
The Open Door (tamper) alarm is initiated  
when the door of the MTEK6000 opens.  
When this occurs, a full wake-up is  
triggered and the MTEK6000 executes its  
processes. The alarm is inactive when the  
door is closed.  
Since there are two individual data banks,  
you may not be able to access all of the  
memory.  
optimized to utilize most or all of the  
memory.  
The configuration must be  
Data may be collected over the telephone  
lines via the modem, on-site through the  
enclosure side connector (RS-232C serial  
port), and/or optionally through the optical  
port (software required for collection). The  
collected data can then be used for:  
Software Error Alarm  
If there is a fault in the software, the  
Software Error alarm will initiate.  
MEMORY (HISTORY LOGGING)  
The MTEK6000 has a total of 96K (Main  
64K bank and Auxiliary 32K bank) of RAM  
for database, audit trail, and history logging.  
With the large memory capacity, over  
81,000 bytes of non-volatile memory is  
reserved for storing historical meter data.  
The non-editable history file provides the  
user with time-related data logged in any  
variation of selectable intervals: minute,  
daily, weekly, and monthly. An event-  
driven history mode allows data logging  
when an event occurs (e.g., alarms). An  
experienced user with software can define  
the type of data and collection period. Since  
history data elements are stored in a memory  
block, the size has to be assigned at the time  
the history process is created in the device,  
typically when the database is downloaded  
at the factory.  
1. Billing information  
2. Measurement reports for utility and  
customer management.  
3. System analysis using flow rate and  
pressure.  
4. Support for estimating gas volume  
consumption in cases of meter or  
instrument malfunction.  
NOTE  
The size of the block CANNOT be changed  
once the history process has been created in  
the unit. A complete download would be  
required to reset the database and change the  
device's memory assignment.  
The total number of records (entries) the  
device can log depends on the number of  
items you wish to log:  
January 2002  
3 -5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
Configuration mode allows you to set-up the  
device's initial configuration, change any of  
the operating parameter values, set alarm  
conditions and limits, or assign function  
Configuration Mode  
Configuration mode allows you to set-up the  
MTEK6000's initial configuration; change  
any of the operating parameter values, and  
keys to a parameter.  
While in the  
configuration mode, the device continues to  
operate normally; it continues to collect  
pulses, sample live pressure and temperature  
values, calculate corrected volume, and  
collect historical data.  
set  
alarm  
conditions  
and  
limits.  
MTEKManager version 1.x or the optional  
external keypad and display are required to  
perform  
configuration.  
See  
the  
MTEKManager on-line help for operating  
instructions. MTEKManager version 1.x is  
supplied, upon request, with your unit  
consisting of Virtual Keypad and other  
utilities.  
Editing Parameters  
1. Enter configuration mode by pressing  
conf (use the password if required).  
2. Display the desired parameter: press  
jump followed by the address of the  
parameter, then press ent (see Appendix  
A for the addresses for your device).  
The function keys can also be used to  
view an assigned parameter.  
Virtual Keypad or Optional External  
Keypad and Display Operation  
To access any operating mode:  
1. Wake-up the device.  
2. Press the keypad key that corresponds to  
the mode you want (Fig. 3-1).  
3. With the desired parameter displayed,  
press edit; the unit will display the  
current parameter value and prompt for a  
new value. Use the keypad to enter the  
correct value and then press ent to  
execute the change. Pressing esc when  
the device prompts for the new value  
will abort the process. Pressing ent if  
nothing has been typed also leaves the  
parameter unchanged.  
conf  
alrm  
cal  
for configuration mode  
for alarm mode  
for calibration mode  
esc to go back to display mode  
While in Display Mode, pressing the and  
keys allow you to scroll through the  
parameter labels. You may use the jump  
key to directly view any of the assigned  
labels; press jump, then enter the label  
number, followed by pressing ent. You can  
use the function keys (F0 - F9) to view  
previously assigned parameters. (software  
lets you assign function keys and labels to  
any parameter.) To enter configuration  
mode, press conf. If configuration mode is  
password protected, the MTEK6000 will  
display ENTER PASSWORD?. Only a  
valid password entry would then be given  
access to this mode.  
3-6  
January 2002  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
Figure 3 - 1 Optional Keypad and Display  
January 2002  
3 -7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
Assigning Function Keys  
Metretek software. Once the audit trail is  
full, the device will not allow any other  
changes to parameters; the audit trail must  
first be uploaded and reset by the host  
software.  
In configuration mode, any parameter can be  
assigned to a function key. To assign a  
function key:  
Wake-Up on Pulse (Event Driven)  
1. Enter configuration mode by pressing  
conf (use the password, if required).  
The MTEK6000 EFC employs a wake-up on  
pulse (event driven wake-up) mode in which  
the unit can be configured to wake-up on a  
specified number of pulses (meter  
revolutions). During wake-up, pressure,  
temperature, and the rest of the analog  
channels are sampled, and the unit executes  
the processes and run calculations.  
2. Display the desired parameter. Press  
jump followed by the address of the  
parameter, and then press ent (see  
Appendix A for parameter addressing  
for your device).  
3. Assign a function key to the parameter.  
Press jump, edit, and then the function  
key you wish to assign to the current  
item [F0 - F9].  
In this mode, the EFC should be configured  
to wake-up on the number of specified  
pulses along with an hourly wake-up to  
record history data. See Appendix A for the  
address of the Wake Up On Pulse (event  
driven) and Wake Up Interval (Seconds)  
parameters.  
NOTE  
There are pre-defined function keys for your  
device - see Appendix A.  
Audit Trail  
Analog Sampling  
The MTEK6000 maintains an electronic  
audit trail file that records all parameter  
changes and calibrations performed on the  
unit. Each entry is identified with the date  
and time the event occurred. The contents  
of this file cannot be changed, providing a  
secure, non-editable audit trail.  
The MTEK6000 has the ability to sample  
the dynamic analog input variable channels  
at intervals from 1-99 seconds.  
enabled, the unit samples pressure,  
temperature, case temperature, and the  
external 1-5 volts or 4-20 ma channels once  
for each selected interval.  
Once  
In the standard MTEK6000 configuration,  
the audit trail is disabled. You can enable  
audit trail logging using Metretek, Inc.  
software or editing the Audit Trail  
Enable/Disable parameter to 250 using the  
keypad and display.  
NOTE  
Battery voltage is NOT sampled at this rate.  
The sample’s minimum and maximum  
values are then checked against their  
corresponding High and Low Setpoints. If  
the device determines these conditions were  
exceeded, it “wakes up” and finds the  
average of all samples accumulated since  
the last process execution, displays this  
analog mean, and applies it to any necessary  
NOTE  
Once the audit trail is enabled, it cannot be  
disabled without downloading  
database with this feature disabled.  
a
new  
Once enabled, the device maintains the audit  
trail file with a maximum of 250 records.  
You can upload the information from the  
unit to a portable or host computer using  
calculations.  
To enable, set Analog  
Sampling Rate parameter to 1-99 seconds.  
3-8  
January 2002  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
Example: A value of 1 will produce a 1 Hz  
(once per second) and a value of 10 a .10 Hz  
(once every 10 seconds) sampling rate. Set  
to 0 to disable.  
Assigning the number of displayed digits  
The number of displayed digits for the  
parameters listed in Appendix A is user-  
configurable. The total number of digits  
before and after the decimal point can be  
from 0 to 8. PcGas Meter Reader Label  
Changer supplied with the Utility Software  
package, or pcGas Host are required to  
change the number of displayed digits. See  
Appendix F for operating instructions on the  
Label Changer.  
NOTE  
Analog sampling will impact battery life in  
battery-operated systems.  
It is only  
recommended for AC or properly sized  
Solar power systems.  
Special Key Combinations  
Viewing and Clearing Alarms  
There are a number of special key  
combinations that allow the user to view  
system information and perform certain  
tasks very easily. They are:  
To enter alarm mode, press alrm. In this  
mode, you can view and acknowledge any  
alarm. If alarms are active, the unit will  
display the first alarm message. If there are  
more alarms, you can view them by pressing  
. Repeatedly pressing cycles through the  
active alarm list.  
F0 and span Displays the MTEK6000 run  
(calculation) time. The unit  
must wake-up by itself at  
least once before a correct  
reading is displayed.  
Alarms can be acknowledged by pressing  
ent while a particular alarm message is  
displayed or by polling with Metretek, Inc.  
software.  
and cal System information (ROM  
version, unit S/N, calculated  
Checksum)  
Unless the parameter's limits are violated  
again, the unit will not include  
acknowledged alarms in its list the next time  
the user enters alarm mode. To exit alarm  
mode without acknowledging the alarm,  
simply press esc.  
and edit Toggles keypad beeper on  
and off  
ent and zero Power-down as soon as  
possible. The unit will not  
power down if the RS-232  
serial cable is connected.  
+/- and zero System Functions (requires  
password)  
January 2002  
3 -9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
CALIBRATION MODE  
1. Display the line pressure by pressing F3  
or jumping to 030302.  
Calibration mode allows the user to calibrate  
any of the analog signals, such as the  
pressure transducer or the temperature  
probe. While operating in the calibration  
mode, the MTEK6000 continues to store  
pulses and periodically updates volume,  
pressure, and temperature data using the  
values measured when calibration mode was  
initially entered. Once in calibration mode,  
the user can perform the following  
operations:  
2. Press cal. Enter your password at the  
optional PASSWORD? prompt, if  
required.  
3. The unit will enter calibration mode.  
The display will alternate between  
CALIBRATING and the parameter label  
(Pressure for example).  
4. Close the shut-off valve between the  
pressure source and the pressure  
transducer.  
1. Calibrate zero only.  
2. Calibrate span only.  
3. Calibrate both zero and span.  
5. Open the pressure sensing line on the  
unit to the atmosphere, and wait until the  
line is fully vented and the reading is  
stable.  
Of course, the option to change the  
calibration reference points is available at all  
times. Several other features make the  
software calibration routine attractive and  
more intuitive. In the MTEK6000, unit  
calibration is software-based; there is no  
need for laborious operator adjustments.  
Software calibration does away with the  
6. Press zero. The display now shows:  
ZERO>  
NEW?>  
XX.XXX  
need  
for  
repetitive  
potentiometer  
adjustments, thereby simplifying field  
calibration procedures.  
XX.XXX represents the unit's default zero  
value. If the current zero reference matches  
the unit's default, simply press ent to collect  
the new point. Otherwise, key in the value  
of the current reference before pressing ent.  
The unit should display Calculating . . .  
briefly, and then display the new point.  
Pressing esc instead of ent at this point  
aborts the operation and returns you to the  
calibration prompt.  
In order to perform calibration, you will  
need a pressure source, temperature source,  
and accurate reference indicators. You will  
also need the Metretek, Inc. Virtual keypad  
or the optional external keypad and display.  
Calibrating the Pressure Transducer  
NOTE  
7. Apply the span (full scale) reference to  
the pressure sensor and wait for the  
reading to stabilize  
Pressing esc repeatedly from anywhere  
within the calibration procedure will back  
the operator out of calibration mode.  
8. Press span. The unit now shows:  
SPAN> XX.XXX  
NEW?>  
3-10  
January 2002  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
As with the zero point, if the external  
reference matches the default span value,  
simply press ent. Otherwise, key in the  
current value of the external reference, then  
press ent. After pressing ent, the display  
reading should immediately adjust to reflect  
the new calibration point. Pressing esc  
instead of ent at this point aborts the  
operation and returns the operator to the  
calibration prompt.  
ZERO> XX.XXX  
NEW?>  
XX.XXX represents the unit's default zero  
value. If the current zero reference matches  
the unit's default, simply press ent to collect  
the new point. Otherwise, key in the value  
of the current reference before pressing ent.  
The unit should display Calculating . . .  
briefly, and then display the new point.  
Pressing esc instead of ent at this point  
aborts the operation and returns you to the  
calibration prompt.  
9. Steps 4 - 8 are required only once. They  
may be repeated as often as necessary  
while in calibration, but only the most  
recent point will be saved on completion  
of calibration.  
6. Place the unit’s temperature probe and  
precision thermometer into a high  
temperature bath. Do not exceed the  
maximum temperature (1700 F). Wait  
for the bath reading to stabilize.  
10. To permanently store the results of the  
calibration press ent, and the unit will  
prompt Enter to accept Calibration.  
Simply press ent again to save the  
calibration. Press esc to abort the  
calibration.  
7. Press span. The unit now shows:  
SPAN> XX.XXX  
NEW?>  
As with the zero point, if the external  
reference matches the default span value,  
simply press ent. Otherwise, key in the  
current value of the external reference, then  
press ent. After pressing ent, the display  
reading should immediately adjust to reflect  
the new calibration point. Pressing esc  
instead of ent at this point aborts the  
operation and returns the operator to the  
calibration prompt.  
Calibrating the Temperature Transducer  
NOTE  
Pressing esc repeatedly from anywhere  
within the calibration procedure will back  
the operator out of calibration mode.  
1. Display the line temperature parameter  
by pressing F4 or jumping to 020304.  
2. Press cal. Enter your password at the  
optional PASSWORD? prompt, if  
required.  
8. Steps 4 - 7 are required only once. They  
may be repeated as often as necessary  
while in calibration, but only the most  
recent point will be saved on completion  
of calibration.  
3. The unit will enter calibration mode.  
(The display will alternate between  
CALIBRATING and the parameter label  
(Flow Temp for example).  
9. To permanently store the results of the  
calibration press ent, and the unit will  
prompt Enter to accept Calibration.  
Simply press ent again to save the  
calibration. Press esc to abort the  
calibration.  
4. Place the unit’s temperature probe into a  
bath of crushed ice. Stir the bath  
continuously and wait for the  
temperature reading to stabilize.  
5. Press zero. The display now shows:  
January 2002  
3 -11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
the differential pressure sensor and wait  
for the reading to stabilize  
Calibrating the Differential Pressure  
Transmitter (EFM Only)  
9. Press span. The unit now shows:  
NOTE  
Pressing esc repeatedly from anywhere  
within the calibration procedure will back  
the operator out of calibration mode.  
SPAN> XX.XXX  
NEW?>  
1. Display the differential pressure by  
As with the zero point, if the external  
reference matches the default span value,  
simply press ent. Otherwise, key in the  
current value of the external reference, then  
press ent. After pressing ent, the display  
reading should immediately adjust to reflect  
the new calibration point. Pressing esc  
instead of ent at this point aborts the  
operation and returns the operator to the  
calibration prompt.  
pressing F8 or jumping to 040302.  
2. Press cal. Enter your password at the  
optional PASSWORD? prompt, if  
required.  
3. The unit will enter calibration mode.  
The display will alternate between  
CALIBRATING and the parameter label  
(Diff Press “H20 for example).  
10. Steps 5 - 9 are required only once. They  
may be repeated as often as necessary  
while in calibration, but only the most  
recent point will be saved on completion  
of calibration.  
4. Open both pressure connection valves.  
5. Open bypass valve and close both high  
pressure and low-pressure transmitter  
connection valves on manifold.  
11. To permanently store the results of the  
calibration press ent, and the unit will  
prompt Enter to accept Calibration.  
Simply press ent again to save the  
calibration. Press esc to abort the  
calibration.  
6. Slowly  
open  
the  
high-pressure  
transmitter connection valve and allow  
the transmitter’s output to stabilize.  
7. Press zero. The display now shows:  
12. Confirm that the bypass valve is open.  
13. Slowly open the high-pressure valve.  
14. Open the low-pressure valve.  
15. Close the bypass valve.  
ZERO>  
NEW?>  
XX.XXX  
XX.XXX represents the unit's default zero  
value. If the current zero reference matches  
the unit's default, simply press ent to collect  
the new point. Otherwise, key in the value  
of the current reference before pressing ent.  
The unit should display Calculating . . .  
briefly, and then display the new point.  
Pressing esc instead of ent at this point  
aborts the operation and returns you to the  
calibration prompt.  
8. Apply the span (full scale) reference to  
3-12  
January 2002  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
CHAPTER 4: OPTIONAL EQUIPMENT  
ANALOG OUTPUT (AO) OPTION  
The Analog Output Module (part # 1021-  
0001B-001) and connecting cable (part #  
1002-0245B-001) provides a two-wire,  
loop-powered, optically isolated, precision  
4-20mA output. This module interfaces  
with the MTEK6000 series product line to  
provide a 4-20mA output for flow rate,  
pressure, or numerous other control and  
monitoring applications. Up to two modules  
can be installed in the MTEK6000 in place  
of the batteries by using the AO module  
adapter plate, (Metretek stk #1008-0011B-  
001).  
Power for the digital interface section of the  
AO is selectable by using the DIPswitches,  
and can be supplied by Vcc of the  
MTEK6000 or from the main supply  
voltage. The AO module interfaces to an  
MTEK6000 series product via the standard  
I2C serial interface bus. The MTEK6000  
unit requests the desired mA output from the  
AO module using this bus. An I2C digital  
I/O chip is used to send the information to  
the analog output section of the AO board.  
The analog output section of the board  
derives its power from the current loop, and  
is optically isolated from the digital control  
interface section. The analog output section  
receives commands from the digital I/O chip  
through opto-isolators. Commands are in  
the form of a serial data stream. The AO  
module provides 4-20mA output signals  
with a resolution of 1 part in 65536 (16 bit).  
The D/A also allows for over-ranging of the  
output to a minimum of 3.5mA and a  
maximum of 24mA. The field interface to  
the D/A is a simple two-wire connection.  
Reverse polarity protection is provided.  
January 2002  
4 - 1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
Figure 4 - 1 Analog Output Option  
4 - 2  
January 2002  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
away with the need for repetitive  
MTEK6000 ANALOG OUTPUT  
SPECIFICATIONS  
potentiometer  
adjustments,  
thereby  
simplifying field calibration procedures.  
The display / keypad or Meter Reader  
Virtual Keypad software are necessary to  
perform software calibration.  
Environmental  
Operating Temperature  
-30°C to + 70°C  
-22°F to 158°F  
0 to 95%  
or  
Operating Humidity  
noncondensing  
NOTE  
Pressing esc repeatedly from anywhere  
within the calibration procedure will back  
the operator out of calibration mode.  
Electrical Isolation  
500 V DC or AC RMS (sine wave) between  
digital interface and 4-20 mA loop.  
1. Display the Analog Output parameter on  
Current Loop Output  
the inside display.  
The common  
Maximum Output Current  
Minimum Output Current  
Maximum Supply Voltage  
Minimum Supply Voltage  
24mA  
3.5mA  
50V  
function key assignment for Analog  
Output #1 is F6, and F7 for Analog  
Output #2.  
8V  
Resolution  
Full Scale %Error  
(software calibrated at  
4 and 20 mA and tested  
at room temperature)  
Temperature Drift  
max  
Error caused by RFI <1% of span shift  
with 2.8W 150MHz  
applied at 1.7’  
16 bits, 0.00024 mA  
±0.01% max  
2. Press cal. Enter your password at the  
optional PASSWORD? prompt, if  
required.  
3. The unit will enter calibration mode and  
the display will show the current value  
and mA representation of the analog  
output signal. For example,  
±0.00044 mA/°F  
Eng: 48.000  
mA: 11.680  
Installing the Analog Output Loop  
where 48.000 is the analog output reading  
representing 11.680 mA. The top line will  
alternate between four different readouts  
Eng: 48.000, UP/DN TO CHANGE,  
CALIBRATING, and the parameter label  
(Analog Output #1 in this case), while the  
bottom line will always show the mA value.  
A ribbon cable connects the AO module to  
the display / analog board. Commands are  
sent to the module via the cable by the  
MTEK6000 device. The 4-20mA current  
loop is a simple two-wire connection. +24  
volts DC nominal is connected to the “+”  
terminal (pos. 1) and the “-”, or return  
terminal (pos. 2), is connected to the field  
instrument to which the 4-20mA signal is  
being sent.  
4. Connect a multimeter in series with the  
loop to measure the current. The field  
instrument which the loop is driving can  
also be used to read the output if desired.  
Calibrating the Analog Output  
5. Pressing will increment the output  
current to represent 0%, 25%, 50%,  
75%, & 100% of the analog output  
parameter to check the calibration.  
Pressing will decrement the output  
current.  
Several features make the Analog Output  
software calibration routine attractive and  
more intuitive. In the MTEK6000 device,  
unit calibration can be software-based; there  
is no need for laborious operator  
adjustments.  
Software calibration does  
January 2002  
4 - 3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
Therefore, 0%  
=
4mA, 25%  
=
8mA, 50% = 12mA, 75% = 16mA, & 100%  
= 20mA.  
x.xxx represents the default span value  
(full scale) and y.yyy is the adjustment  
made to 20mA for the analog output  
signal. The adjustment can either be  
positive or negative shown by + or -  
respectively. The top line will alternate  
between span: x.xxx and UP/DN TO  
ADJUST. This is the span adjusted  
value to calibrate the analog output to  
20mA.  
6. If adjustments are needed, press zero.  
The display now shows:  
zero:  
x.xxx  
04.000 +y.yyy mA  
x.xxx represents the default zero value  
(low scale) and y.yyy is the adjustment  
made to 4mA for the analog output signal.  
The adjustment can either be positive or  
negative shown by + or - respectively. The  
top line will alternate between zero: x.xxx  
and UP/DN TO ADJUST. This is the zero  
adjusted value to calibrate the analog output  
to 4mA.  
9. Press or to increase or decrease the  
output until the meter reads 20mA or the  
current span reference matches the field  
equipment. Press ent to collect the new  
point.  
10. Steps 5 - 9 are required only once. They  
may be repeated as often as necessary  
while in calibration, but only the most  
recent point will be saved on completion  
of calibration.  
7. Press or to increase or decrease the  
output until the meter reads 4mA or the  
current zero reference matches the field  
equipment. Press ent to collect the new  
point.  
11. To permanently store the results of the  
calibration press ent, and the unit will  
prompt Enter to accept Calibration.  
Simply press ent again to save the  
8. Press span. The display shows:  
calibration.  
calibration.  
Press esc to abort the  
span:  
20.0  
x.xxx  
+y.yyy mA  
4 - 4  
January 2002  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
Changing the Battery  
Chapter 5 : MAINTENANCE and  
SOFTWARE PACKAGES  
To replace the battery in the unit:  
As with any device based on solid-state  
electronics, actual maintenance of the  
MTEK6000 should be minimal. However,  
there are certain guidelines that, if followed,  
will minimize device failure and increase  
the product’s service life.  
1. Open the front door by loosening the  
upper and lower right hand corner  
screws of the device enclosure and  
swinging the door out (see Fig 1-1 and  
1-2 in Chapter 1).  
2. Attach the new battery to connector J-6  
or J-7 (see Fig. 2-1 in Chapter 2).  
Enclosure Maintenance  
Enclosure maintenance is a program of  
routine inspections to insure the integrity of  
the lid's seal and the various ports in the  
box's exterior. Excess moisture can ruin an  
MTEK6000 if allowed to accumulate within  
the enclosure. Although the circuit boards  
themselves are conformal coated to protect  
3. Disconnect the old battery from the  
other connector in the unit.  
4. Press any key to wake-up the unit and  
verify that it is fully operational.  
Calibration  
against  
humidity,  
the  
wiring  
interconnections and various exposed metal  
surfaces are susceptible to corrosion in  
extreme cases of interior humidity. Here are  
some checks you should periodically make  
of the enclosure:  
Calibration is a crucial element of any  
scheduled maintenance program. However,  
because of the unit’s design, software  
calibration does away with the need for  
laborious adjustments, simplifying field  
calibration. See the section on Calibration  
for more details.  
1. Ensure the unit’s mounting arrangement  
is secure and provides a stable platform  
for termination of the pressure tubing.  
2. Verify the integrity of the enclosure lid  
seal.  
Check the lid gasket for  
deterioration, chemical damage, tears, or  
compression.  
3. Check for damaged cord grips and  
missing or damaged RS-232C port caps.  
Liquid must not be allowed to  
accumulate within the interior of the  
enclosure.  
4. Examine the RS-232C port. Ensure the  
port's mounting screws are secure and  
provide firm support when attaching a  
serial cable.  
January 2002  
5 - 1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
Label Changer - View and change  
label and function key definitions.  
Software  
Important Note: pcGas Meter Reader,  
pcGas Host and pcGas Customer Monitor  
applications are DOS based programs.  
They are available for a one-time charge  
but are sold ‘as is’ and are not being  
changed or upgraded in any way by  
Metretek. While these programs may be  
of value to certain users, Metretek makes  
no warranty as to their performance.  
Metretek strongly encourages the use of  
the MTEKManager and DC2000 32 bit  
Windowstm applications.  
Virtual Keypad -  
Emulates the  
keypad in the MTEK6000 products. Can  
be used for configuration and  
calibration.  
pcGas Customer Monitor  
pcGas Customer Monitor lets personnel  
responsible for collecting data to  
conveniently interact with the unit. No data  
can be modified in the unit with this  
software - it only provides a convenient  
method of viewing and reading history data  
stored in the unit.  
pcGas Meter Reader Software  
pcGas Meter Reader is a flexible, yet simple  
software package that allows personnel  
responsible for site-specific configuration or  
data collection to conveniently interact with  
the unit. pcGas Meter Reader can be  
purchased to interface with the unit, but  
does not replace pcGas Host software; a user  
cannot use it to download processes to an  
MTEK6000 with blank system memory.  
However, it does allow a convenient method  
of viewing and/or modifying general site-  
specific database information (most of  
which are shown in Appendix A).  
MTEK Manager  
The MTEK Manager software is an  
integrated group of utilities designed for  
configuration and management of the  
MTEK6000 corrector as well as the  
AE5000/6000 product lines. The software  
utilizes MS Access™ compatible databases  
for all of its data, and is suitable for  
managing small groups of correctors.  
Mtek Manager is licensed for end-use  
pursuant to Metretek’s standard licensing  
fees and terms. This package has all the  
features of the basic package, but adds the  
ability to retrieve, view, and print the  
historical data from the correctors.  
It also adds the ability to perform remote  
access using a dialup phone system.  
Included is an auto-polling application that  
can be used to schedule polls to the  
Trend Graphics, and AutoPoll are standard  
features with pcGas Meter Reader. Manuals  
for Meter Reader and Meter Reader Utilities  
are provided with each registered copy of  
the program. Refer to these manuals for  
additional information.  
pcGas Meter Utility Package  
configured stations, as well as answer  
incoming calls. Data export to DC2000™  
can also be automated after data is collected.  
The pcGas Meter Reader Utility Package is  
supplied, upon request, with your unit. See  
Appendix for installation and operating  
instructions. The utility package consists of  
the following:  
Site I.D. Changer - View or change  
the Site I.D. stored in the device.  
5 - 2  
January 2002  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
DC2000  
The MTEK6000 is fully compatible with  
Metretek’s DC2000. DC2000 is Metretek’s  
flagship collection and management  
software system for energy data. DC2000’s  
scaleability and flexibility enables users to  
choose from a wide range of functions and  
data throughput configurations. This  
protects your investment by letting you  
continuously adapt your system to operate in  
proportion to your business needs. See your  
authorized Metretek representative for  
complete details on DC2000 capabilities and  
licensing terms.  
January 2002  
5 - 3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
5 - 4  
January 2002  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
APPENDIX A: PROCESS CONFIGURATION STANDARD  
The MTEK6000 uses Process configuration for database organization and management.  
Table A-1: Standard display mode and  
function keys for MTEK6000 EFC  
Table A-2: Standard alarms for  
MTEK6000 EFC  
Alarms  
Alarm Code  
Label I.D.  
Description  
Address  
First Time Power  
Low Supply Volts  
High Flow Rate  
Low Flow Rate  
High Pressure  
130401  
130501  
130601  
130701  
140401  
140501  
140601  
140701  
150401  
150501  
150601  
150701  
160401  
160501  
160601  
160701  
170401  
F1 1 CV  
F2 2 UV  
F3 3 PR  
F4 4 FT  
5 FR  
Corr Volume MCF 051102  
Uncor Volume MCF 051108  
PRessure PSG  
Flow Temp F  
Flow Rate MCH 050302  
Supply Volts VDC 020309  
Case Temp F  
Flow Constant  
Prev Day vol MCF 050903  
Curr Day vol MCF 050905  
Cubic Unit/p CFP  
Press Scale PSG  
030302  
020304  
Low Pressure  
Current Day Flow  
Faulty Counter  
*High Temperature  
*Low Temperature  
Lost Pressure  
*Lost Temperature  
LowVolt Shutdown  
Open Door  
F5 6 SV  
7 CT  
020302  
050601  
8 FL  
9 PD  
10 CD  
11 CU  
050802  
020402  
12 PS  
Software Error  
Switch #1 Alarm  
Switch #2 Alarm  
*MTEK6000  
EFCV  
only  
Table A-3: Standard history data stored  
in the MTEK6000 EFC and EFCP  
40 days of daily corrected volume  
40 days of daily uncorrected volume  
40 days of daily maximum flow rate  
40 days of daily minimum flow rate  
40 days of daily average pressure  
40 days of daily average temperature  
40 days of hourly corrected volume  
40 days of hourly uncorrected volume  
40 days of hourly average pressure  
40 days of hourly average temperature  
40 days of hourly instantaneous supply  
voltage (snapshots).  
40 days of hourly case temperature  
(snapshots)  
January 2002  
A-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
Table A-4: Standard display mode and  
function keys for MTEK6000 EFC w/  
Aux Pressure  
Table A-6: Standard history data stored  
in the MTEK6000 EFC w/ Aux Pressure  
40 days of daily corrected volume  
40 days of daily uncorrected volume  
40 days of daily maximum flow rate  
40 days of daily minimum flow rate  
40 days of daily average pressure  
40 days of daily average aux. pressure 1  
40 days of hourly average aux. pressure 1  
40 days of hourly uncorrected volume  
40 days of hourly average pressure  
40 days of hourly average temperature  
40 days of hourly instantaneous supply  
voltage (snapshots).  
Label I.D.  
Description  
Address  
F1 1 CV  
F2 2 UV  
F3 3 PR  
F4 4 FT  
5 FR  
F6 6 AP  
F5 7 SV  
8 CT  
Corr Volume MCF 051102  
Uncor Volume MCF 051108  
PRessure PSG  
Flow Temp F  
Flow Rate MCH 050302  
Aux Pressure PSG 020305  
Supply Volts VDC 020309  
030302  
020304  
Case Temp F  
Flow Constant  
020302  
050601  
9 FL  
40 days of hourly case temperature  
(snapshots)  
10 PD  
11 CD  
Prev Day vol MCF 050903  
Curr Day vol MCF 050905  
12 CU  
13 PS  
14 AS  
Cubic Unit/p CFP  
Press Scale PSG  
Aux p Scale PSG  
050802  
020402  
020404  
Table A-5: Standard alarms for  
MTEK6000 EFC w/ Aux Pressure  
Alarms  
Alarm Code  
First Time Power  
Low Supply Volts  
High Flow Rate  
Low Flow Rate  
High Pressure  
160401  
160501  
160601  
160701  
170401  
170501  
170601  
180401  
180501  
180601  
180701  
170701  
200401  
200501  
200601  
190401  
190501  
190601  
Low Pressure  
Current Day Flow  
High Temperature  
Low Temperature  
Lost Pressure  
Lost Temperature  
Faulty Counter  
LowVolt Shutdown  
Open Door  
Software Error  
High Aux Press  
Low Aux Press  
Lost Aux Press  
A-2  
January 2002  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
Table A-7: Standard display mode and  
function keys for MTEK6000 EFC w/2  
Aux Pressure  
Table A-8: Standard alarms for  
MTEK6000 EFCV w/2 Aux Pressure  
Alarms  
Alarm Code  
Label I.D.  
Description  
Address  
First Time Power  
Low Supply Volts  
High Flow Rate  
Low Flow Rate  
High Pressure  
190401  
190501  
190601  
190701  
200401  
200501  
200601  
210401  
210501  
210601  
210701  
200701  
230401  
230501  
230601  
220401  
220501  
220601  
240401  
240501  
240601  
230701  
240701  
F1 1 CV  
F2 2 UV  
F3 3 PR  
F4 4 FT  
F6 5 A1  
F7 6 A2  
F5 7 SV  
8 FR  
Corr Volume MCF 051102  
Uncor Volume MCF 051108  
PRessure PSG  
030302  
020304  
020305  
020306  
Flow Temp deg F  
Aux 1 press PSG  
Aux 2 press PSG  
Low Pressure  
Current Day Flow  
High Temperature  
Low Temperature  
Lost Pressure  
Lost Temperature  
Faulty Counter  
LowVolt Shutdown  
Open Door  
Supply Volts VDC 020309  
Flow Rate MCH 050302  
Uncorr Flow MCH 050306  
9 UF  
10 CT  
11 FC  
12 PD  
13 CD  
14 CU  
15 PS  
16 AS  
Case Temp F  
Flow Constant  
020302  
050601  
Prev Day vol MCF 050903  
Curr Day vol MCF 050905  
Cubic Unit/p CFP  
Press Scale PSG  
Aux 1p Scale psg  
Aux 2p Scale psg  
050802  
020402  
020404  
020405  
Software Error  
High Aux 1 Press  
Low Aux 1 Press  
Lost Aux 1 Press  
High Aux 2 Press  
Low Aux 2 Press  
Lost Aux 1 Press  
Switch #1 Alarm  
Switch #2 Alarm  
17 AS  
January 2002  
A-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
Table A-9: Standard history data stored  
in the MTEK6000 EFC w/ 2 Aux  
Pressure  
Table A-10: Standard display mode and  
function keys for MTEK6000 EFC2 w/2  
Aux Pressure  
Label I.D.  
Description  
Address  
40 days of daily corrected volume  
40 days of daily uncorrected volume  
40 days of daily maximum flow rate  
40 days of daily minimum flow rate  
40 days of daily average pressure  
F1 1 CV  
F2 2 UV  
F3 3 PR  
F4 4 FT  
5 F1  
Corr Volume MCF 051102  
Uncor Volume MCF 051108  
PRessure PSG  
030302  
020304  
050302  
050306  
050601  
050903  
050905  
050802  
061102  
Flow Temp deg F  
Flow rate 1 MCH  
Uncorr Flow R1  
Flow Constant 1  
Prev Day Vol 1  
Curr Day Vol 1  
Cubic Unit/p 1  
Corr vol 2 MCF  
40 days of daily maximum pressure  
40 days of daily minimum pressure  
40 days of daily average aux. pressure 1  
40 days of daily average aux. pressure 2  
40 days of daily maximum aux. pressure 1  
40 days of daily maximum aux. pressure 2  
40 days of daily minimum aux. pressure 1  
40 days of daily minimum aux. pressure 2  
40 days of hourly average aux. pressure 1  
40 days of hourly average aux. pressure 2  
40 days of hourly uncorrected volume  
40 days of hourly average pressure  
6 UF  
7 FC  
8 PD  
9 CD  
10 CU  
11 C2  
12 U2  
13 F2  
14 UF  
15 FC  
16 PD  
17 CD  
18 CU  
19 A1  
20 A2  
21 SV  
22 CT  
23 PR  
24 A1  
25 A2  
37 TF  
Uncorr vol 2 MCF 061108  
Flow rate 2 MCH  
Uncorr Flow R2  
Flow Constant 2  
Prev Day Vol 2  
Curr Day Vol 2  
Cubic Unit/p 2  
060302  
060306  
060601  
060903  
060905  
060802  
020305  
020306  
Aux 1 press PSG  
Aux 2 press PSG  
Supply Volts VDC 020309  
Case Temp F  
020302  
020402  
020404  
020405  
370201  
PRessure SCL  
Aux press 1 SCL  
Aux press 2 SCL  
Total Flow Rate  
A-4  
January 2002  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
Table A-11: Standard alarms for  
MTEK6000 EFC2 w/2 Aux Press  
Table A-12: Standard history data  
stored in the MTEK6000 EFC2 w/ 2 Aux  
Press  
Alarms  
Alarm Code  
40 days of daily corrected volume 1  
40 days of daily corrected volume 2  
40 days of daily uncorrected volume 1  
40 days of daily uncorrected volume 2  
40 days of daily maximum flow rate 1  
40 days of daily maximum flow rate 2  
40 days of daily minimum flow rate 1  
40 days of daily minimum flow rate 2  
40 days of daily average pressure  
40 days of daily average aux. pressure 1  
40 days of daily average aux. pressure 2  
40 days of daily temperature  
37 days of hourly corrected volume 1  
37 days of hourly corrected volume 2  
37 days of hourly uncorrected volume 1  
37 days of hourly uncorrected volume 2  
37 days of hourly average pressure  
37 days of hourly aux. pressure 1  
37 days of hourly aux. pressure 2  
37 days of hourly temperature  
First Time Power  
Low Supply Volts  
LowVolt Shutdown  
Software Error  
210401  
210501  
210601  
210701  
220401  
220501  
220601  
220701  
230401  
230501  
230601  
230701  
240401  
240501  
240601  
240701  
250401  
250501  
250601  
250701  
260401  
260501  
260601  
260701  
270401  
270501  
270601  
Lost Pressure  
Lost Temperature  
Lost Aux 1 Press  
Lost Aux 2 Press  
High Flow Rate 1  
Low Flow Rate 1  
Curr Day Volume 1  
Faulty Counter 1  
High Flow Rate 2  
Low Flow Rate 2  
Cur Day Volume 2  
Faulty Counter 2  
High Pressure  
Low Pressure  
High Temperature  
Low Temperature  
High Aux 1 Press  
Low Aux 1 Press  
High Aux 2 Press  
Low Aux 2 Press  
Open Door  
37 days of hourly instantaneous supply  
voltage (snapshots).  
37 days of hourly case temperature  
(snapshots)  
Switch 1 Alarm  
Switch 2 Alarm  
January 2002  
A-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
Table A-13: Standard display mode and  
function keys for MTEK6000 EFM w/  
Aux Press  
Table A-14: Standard alarms for  
MTEK6000 EFM w/ Aux Press  
Alarms  
Alarm Code  
Label I.D.  
Description  
Address  
First Time Power  
Low Supply Volts  
High Flow Rate  
Low Flow Rate  
High Pressure  
180401  
180501  
180601  
180701  
190401  
190501  
190601  
190701  
200401  
200501  
200601  
210401  
210501  
210601  
210701  
220401  
220501  
220601  
220701  
200701  
F1 1 CV  
F8 2 DP  
F3 3 PR  
F4 4 FT  
5 FR  
F6 6 AP  
7 OD  
F5 8 SV  
9 CT  
Corr Volume MCF 051102  
Diff Press “H20  
PRessure PSG  
Flow Temp F  
040302  
030302  
020304  
050302  
Flow Rate MCH  
Low Pressure  
Aux Pressure PSG 020306  
Orif Diameter IN 050608  
High Diff Press  
Low Diff Press  
High Temperature  
Low Temperature  
Current Day Flow  
Lost Pressure  
Lost Temperature  
Lost Diff Press  
LowVolt Shutdown  
Open Door  
Supply Volts DC  
Case Temp F  
Flow Constant  
020309  
020302  
050601  
10 FC  
11 PD  
12 CD  
13 PS  
14 DS  
15 AS  
Prev Day vol MCF 050903  
Curr Day vol MCF 050905  
Press Scale PSG  
Dp Scale H20  
Aux p Scale PSG  
020402  
020404  
020405  
Software Error  
High Aux Press  
Low Aux Press  
Lost Aux Press  
Table A-15: Standard history data  
stored in the MTEK6000 EFM w/ Aux  
Pressure  
40 days of daily corrected volume  
40 days of daily average temperature  
40 days of daily maximum flow rate  
40 days of daily minimum flow rate  
40 days of daily average pressure  
40 days of daily average aux. pressure  
40 days of daily average differential pressure  
40 days of hourly corrected volume  
40 days of hourly average differential  
pressure  
40 days of hourly average pressure  
40 days of hourly average aux. pressure  
40 days of hourly average temperature  
40 days of hourly instantaneous supply  
voltage (snapshots).  
40 days of hourly case temperature  
(snapshots)  
A-6  
January 2002  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
Table A-18: Standard history data  
stored in the MTEK6000 EPR  
Table A-16: Standard display mode and  
function keys for MTEK6000 EPR  
40 days of daily average pressure  
40 days of daily maximum pressure  
40 days of daily minimum pressure  
40 days of hourly average pressure  
40 days of hourly instantaneous supply  
voltage (snapshots).  
40 days of hourly case temperature  
(snapshots)  
15 days of 10-minute pressure (snapshots)  
Label I.D.  
Description  
Address  
F1 1 P1  
2 PS  
Pressure #1  
030302  
020402  
080605  
080606  
080705  
080706  
020309  
020302  
Pressure Scale  
High press In al  
High press Out al  
Low press In al  
Low press Out al  
Supply Voltage  
Case Temperature  
3 HI  
4 HO  
5 LI  
6 LO  
F6 7 SV  
F2 8 CT  
F3 9 DO  
F4 10 SO  
DOOR 0=open 1=cl 070302  
SWITCH 0=OPEN 070402  
11 AW Alarm Wakeup INT 170610  
Table A-17: Standard alarms for  
MTEK6000 EPR  
Alarms  
Alarm Code  
First Time Power  
Low Supply Volts  
High Pressure  
Low Pressure  
Lost Pressure  
LowVolt Shutdown  
Open Door  
080401  
080501  
080601  
080701  
090401  
090501  
090601  
090701  
100401  
Switch Alarm  
Reset Min. Hist  
January 2002  
A-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
Table A-19: Standard display mode and  
function keys for MTEK6000 ETR  
Table A-21: Standard history data  
stored in the MTEK6000 ETR  
40 days of daily average temperature  
40 days of daily maximum temperature  
40 days of daily minimum temperature  
40 days of hourly average temperature  
40 days of hourly instantaneous supply  
voltage (snapshots).  
40 days of hourly case temperature  
(snapshots)  
15 days of 10-minute temperature  
(snapshots)  
Label I.D.  
Description  
Address  
F4 1 FT  
2 HI  
Flow Temp F  
020304  
080605  
080606  
080705  
080706  
020309  
020302  
160302  
160402  
160403  
170302  
170402  
170403  
070302  
070401  
070402  
High temp In al  
High temp Out al  
Low temp In al  
Low temp Out al  
Supply Volts DC  
Case Temp F  
Analog Output #1  
AOut Hi Scale #1  
AOut Lo Scale #1  
Analog Output #2  
AOut Hi Scale #2  
AOut Lo Scale #2  
Open Door status  
Status input #1  
Status input #2  
3 HO  
4 LI  
5 LO  
F5 6 SV  
7 CT  
F6 8 AO  
9 AO  
10 AO  
F7 11 AO  
12 AO  
13 AO  
14 OD  
15 S1  
16 S2  
Table A-20: Standard alarms for  
MTEK6000 ETR  
Alarms  
Alarm Code  
First Time Power  
Low Supply Volts  
High Temperature  
Low Temperature  
Lost Temperature  
LowVolt Shutdown  
Open Door  
080401  
080501  
080601  
080701  
090401  
090501  
090601  
090701  
Software Error  
A-8  
January 2002  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
Table A-22: Standard display mode and  
function keys for MTEK6000 EPTR  
Table A-23: Standard alarms for  
MTEK6000 EPTR  
Label I.D.  
Description  
Address  
Alarms  
Alarm Code  
F3 1 PR  
2 PS  
PRessure PSG  
030302  
020402  
140405  
140406  
140505  
140506  
020304  
140605  
140606  
140705  
140706  
020309  
020302  
220302  
220402  
220403  
230302  
230402  
230403  
120302  
120401  
120402  
First Time Power  
Low Supply Volts  
High Pressure  
130401  
130501  
130601  
130701  
140401  
140501  
140601  
140701  
150401  
150501  
150601  
Press Scale PSG  
High press In al  
High press Out al  
Low press In al  
Low press Out al  
Flow Temp F  
High temp In al  
High temp Out al  
Low temp In al  
Low temp Out al  
Supply Volts DC  
Case Temp F  
Analog Output #1  
AOut Hi Scale #1  
AOut Lo Scale #1  
Analog Output #2  
AOut Hi Scale #2  
AOut Lo Scale #2  
Open Door status  
Status input #1  
Status input #2  
3 HI  
4 HO  
5 LI  
6 LO  
Low Pressure  
High Temperature  
Low Temperature  
Lost Pressure  
Lost Temperature  
LowVolt Shutdown  
Open Door  
F4 7 FT  
8 HI  
9 HO  
10 LI  
11 LO  
F5 12 SV  
13 CT  
F6 14 AO  
15 AO  
16 AO  
F7 17 AO  
18 AO  
19 AO  
20 OD  
21 S1  
Software Error  
Table A-24: Standard history data  
stored in the MTEK6000 EPTR  
40 days of daily average pressure  
40 days of daily maximum pressure  
40 days of daily minimum pressure  
40 days of daily average temperature  
40 days of daily maximum temperature  
40 days of daily minimum temperature  
40 days of hourly average pressure  
40 days of hourly average temperature  
40 days of hourly instantaneous supply  
voltage (snapshots).  
22 S2  
40 days of hourly case temperature  
(snapshots)  
15 days of 10-minute pressure (snapshots)  
January 2002  
A-9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
A-10  
January 2002  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
AGA-8 Supercompressibility Gross Method 1  
APPENDIX B: CALCULATIONS  
The EFC is configured with values for BTU  
content, specific gravity and mol percent of  
CO2.  
AGA-7 Volume Calculations  
The MTEK6000 EFC performs volume  
calculations based on the Ideal Gas Law.  
Boyle’s Law is used for pressure and  
Charles’s Law for temperature. These laws  
state that the volume of any definite weight of  
a perfect gas varies inversely with change in  
absolute pressure and directly with change in  
absolute temperature. The unit can perform  
such calculations for turbine, rotary, and  
diaphragm displacement meters.  
AGA-8 Supercompressibility Gross Method 2  
The EFC is configured with values for  
specific gravity and mol percent of N2 and  
CO2.  
Vc = Vu x (Pm + Pa) x Tb + 459.67 x (Fpv)2  
Pb  
Tm + 459.67  
Where:  
Vc = Volume corrected to base conditions  
Vu = Uncorrected line volume  
Pm = Measured line pressure (psig)  
Pa = Atmospheric pressure  
Pb = Base pressure  
Tb = Base temperature (0F)  
Tm = Measured line temperature (0F)  
Fpv = Supercompressibility factor  
The unit makes continuous correction for the  
specified gas composition based on actual  
sensed pressure and temperature.  
NOTE  
All of the AGA factors can be based on either  
flowing or constant values.  
The supercompressibility factor can be  
calculated from either NX-19 or AGA-8  
Gross Methods 1 and 2 reports and applied to  
the volume equation.  
NX19 Supercompressibility Report  
The EFC is configured with values for  
specific gravity, mol percent of N2 and CO2.  
January 2002  
B-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
B -2  
January 2000  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
To enable analog sampling, set this parameter  
to 1-99 seconds. Set to 0 to disable analog  
sampling.  
APPENDIX C:  
PARAMETER DESCRIPTION  
The parameters relative to the operation and  
configuration of the MTEK6000 are listed  
below (See Appendix A for the addresses of  
these parameters).  
Atmospheric Pressure  
If the station is configured for ABSOLUTE  
pressure, enter 0.  
atmospheric pressure for GAUGE pressure.  
The default value is 14.4 PSI for gauge and 0  
for absolute.  
Enter the actual  
Alarm Pulse Output Enable/Disable  
The EFC can generate a generic pulse output  
on any alarm condition. This parameter  
enables the pulse output through Relay #2.  
Enter 35.7 to enable or 0.0 to disable. The  
Default value is 0.  
Audit Trail Logging Enable  
This parameter enables Audit Trail logging in  
the MTEK6000. Examples of audit trail  
events include editing any item at the device,  
and logging the time it enters calibration  
mode. After the maximum number of events  
(250) has been logged, no more parameters  
can be changed in the unit until the audit trail  
is uploaded and reset by the software. This  
parameter is disabled at the factory. The user  
may enable Audit Trail logging by entering  
250. Default value is 0 for disable.  
Alarm Pulse Time (ms)  
This parameter sets the output band width in  
milliseconds (ms) for pulse outputs generated  
by an alarm condition. The user should take  
care in assigning a value for this parameter.  
The value should always be less than the  
Wakeup Interval Parameter; else the unit will  
stay awake for the duration of the pulse.  
Range 0 to 32,000 ms. The Default value is  
70.  
NOTE  
Once the audit trail is enabled, the user cannot  
disable it without reloading the database.  
Analog Sampling Rate (seconds)  
This parameter sets the rate at which the unit  
will sample its dynamic analog input variable  
channels. When enabled, the unit samples  
pressure, flow temperature, case temperature  
and the external analog transmitters values  
once for each selected interval.  
Base Pressure  
The Base Pressure parameter appears as a  
factor in the Corrected Volume calculation. It  
is one of the factors used to correct the  
flowing volume, as registered by the meter  
itself, to the base volume used for calculating  
“Standard Volume”. The normal pressure  
ranges for this parameter are 11.000 to 16.000  
PSI.  
The sample’s minimum and maximum values  
are then checked against their corresponding  
High and Low Setpoints.  
If the unit  
determines that these conditions were  
exceeded, it “wakes up” and finds the average  
of all samples accumulated since the last  
process execution, displays this analog mean,  
and applies it to any necessary subsequent  
calculation.  
The standard value for this parameter is  
14.730 PSIA, 101.560 kPa, or 1.114 kg/cm2,  
depending upon the system of units chosen  
for the particular device. Default value is  
14.73.  
January 2002  
C - 1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
Base Temperature  
Corrected Volume  
The Base Temperature parameter appears as a  
factor in the Corrected Volume calculation. It  
is one of the factors used to correct the  
flowing volume, as registered by the meter  
itself, to the base volume used for calculating  
“Standard Volume”. The default value of this  
parameter is 60.000 degrees F.  
The corrected volume is calculated based  
upon AGA report #7 or #8 and reflects the  
corrected volumetric flow taking the base  
conditions into consideration. The Initial  
value is 0.  
Corrected Volume (Pressure Only)  
The corrected volume is calculated based  
upon AGA report #7 and reflects the  
corrected volumetric flow taking only the  
Pressure base conditions into consideration.  
The temperature factors have no effect. The  
Initial value is 0.  
Calibrate Mode Time-out  
The fractional portion of this parameter is the  
Calibrate Mode Time-out. It controls how  
long the unit will remain powered up in the  
calibration mode if no key is being pressed.  
For example, a calibration time-out of 30  
minutes would be entered as XX.30. The  
default value is 60.30.  
Corrected Volume and Flow Multiplier  
This multiplier specifies the output value for  
corrected volume and flow. For example, if  
the device is to measure flow in thousands of  
cubic feet (MCF), select Thousands (1000's)  
of units from the pick list. The Default value  
is 1000.  
Calibration Password  
A password can be entered to prevent  
unauthorized users from accessing calibration  
mode. The value may be up to six digits long  
(including the decimal point and sign). The  
Default value is 0.  
Ten Thousands (10,000's) of units 10000  
Thousands (1000's) of units  
Hundreds (100's) of units  
Tens (10's) of units  
1000  
100  
10  
Configuration Password  
A password can be entered to prevent  
unauthorized  
users  
from  
accessing  
Single (1's) units  
1
configuration mode. The value may be up to  
six digits long (including the decimal point  
and sign). The Default value is 0.  
Counter Fault Monitoring (EFC Only)  
This parameter can be used to enable or  
disable fault monitoring on the main counter  
#1. Form C (3-wire) input is required for this  
function. In this mode, two working counters  
are monitored for discrepancies. A “Delta”  
value will be accumulated which reflects any  
discrepancies between the counters. If any of  
the dual-reed switches should be defective,  
and the “Delta” exceeds the Counter Fault  
Threshold parameter, the input pulses will  
automatically switch to either working single-  
ended counter. The EFC will generate a  
Faulty Counter alarm. To enable, set this  
parameter to 1 and to disable set to 0. The  
default value is 0 for disable.  
Corrected Pulse Output -Cubic Unit / Pulse  
This parameter should reflect the value of the  
corrected pulse output sent from the station to  
an external device. For example, if each  
output pulse represents 1000 cubic feet, then  
this parameter should be set to 1000. The  
Default value is 1000.  
Corrected Pulse Out Enable/Disable  
Corrected pulses can be sent to an external  
device by enabling the Pulse Output Channel.  
The rate at which pulses are generated is  
determined by the Corrected Pulse Out - CF  
per Pulse parameter. To enable Relay #1 for  
pulse output, enter a value of 35.8. Enter -1  
to disable. The Default value is -1.  
C - 2  
January 2002  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
Cubic Unit/Pulse In or Meter Drive  
Date (Month, Day, Year)  
This parameter determines the volume unit  
represented by one input pulse, and is  
normally set to equal the drive (CF/Rev) of  
the meter. Standard indexes produce one  
pulse per revolution; therefore, the CF/Pulse  
will equal the drive rate of the meter. The  
Default value is 100.  
This is the current Date in the unit  
(MMDDYY). It is updated on each process  
scan.  
Flow Rate Update Interval (s)  
This parameter determines how often the flow  
rate gets updated in the EFC. However, it has  
no control over Volume updates. Volume is  
updated every time the unit wakes up from  
the sleep mode or every 3-4 seconds if the  
unit is always awake.  
For instruments utilizing remote pulsing  
devices, this value can also be determined by  
dividing the CF/Rev of the meter by the  
number of pulses per revolution generated by  
the remote pulser.  
The Flow Rate Update Interval parameter is  
of the form XXX.YY.M where the integer  
portion (XXX) is the time interval in seconds  
between Flow Rate updates. The digits after  
the decimal point (YY.M) represent the zero-  
flow period (i.e. the amount of time to wait  
without receiving pulses before generating a  
zero flow rate). The first two digits are  
minutes and the third digit (T) is a fraction of  
a minute. Thus, a two-minute zero-flow  
period would be entered as XXX.020, a ten  
minute period as XXX.10. The maximum  
allowed period is 27 minutes and the default  
is one minute (XXX.010).  
For example: Consider a 3GT meter with a  
100 CF/Rev drive and a 50 Pulse/Rev Imac  
Pulsamatic Transmitter. The CF/Pulse is  
determined as follows:  
CF/Pulse = 100 CF/Rev * 1 Rev/50 Pulses  
= 2 CF/Pulse  
NOTE FOR METERS WITH 5 CF/REV  
DRIVE RATES:  
Re-position the input compound gear to the  
correct position, and set this parameter to a  
value of 10. Refer to Section Two or contact  
Metretek, Inc. for assistance in positioning the  
compound gear.  
The user MUST specify a zero-flow period,  
which is larger than or equal to the Flow Rate  
Update Interval. Specifying a zero-flow  
period smaller than the Flow Rate Update  
Interval will result in a zero-flow rate being  
displayed even when there is an apparent  
flow. For example: A value of 600.11 would  
update flow rate every 10 minutes and display  
zero flow if 11 minutes passed without an  
input pulse. A value of 600.01 will not work  
properly. It would try to update flow rate  
every 10 minutes, but would zero the flow  
rate every minute, even if pulses come in  
faster than one per minute.  
Current Day High Volume Alarm Reset  
After the Current Day's Total is exceeded, the  
setpoint at which the unit exits this condition  
is entered in this location. The Default value  
is 99,990.  
Current Day High Volume Alarm  
Setpoint  
This parameter specifies the setpoint at which  
the unit determines whether the Current Day  
Flow Total is in an alarm condition. For  
transport or interruptible customers, this  
parameter can be used to alarm when an  
account has exceeded a predetermined daily  
volume allocation. The Default value is  
100,000.  
January 2002  
C - 3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
Flow Units  
High Pressure Alarm Reset  
The flow units parameter reflects the time  
used to represent the flow rate. For example,  
if the flow rate represents cubic feet per hour,  
this parameter should be set to Hour. The  
Default value is 2.  
After a High Pressure alarm occurs, the  
setpoint at which the unit exits this condition  
is entered in this location. The Default value  
is 1480.  
High Temperature Alarm Reset  
After a High Temperature alarm occurs, the  
setpoint at which the unit exits this condition  
is entered in this location. The Default value  
is 180.  
Minute  
Hour  
Day  
1
2
3
Gas Day Roll Time HHMM (Hours,  
Minutes)  
High Temperature Alarm Setpoint  
The setpoint at which the unit determines that  
there is a High Temperature alarm condition.  
The Default value is 200.  
This item is used in Daily and Monthly  
history modes to determine when the gas day  
ends. The time is entered in military time.  
For example, a standard roll time of 8:00AM  
is entered as 800.0. The Range is 0 to 2359.  
The default value is 800.  
Low Differential Pressure Alarm Reset  
After a Low Differential Pressure alarm  
occurs, the setpoint at which the unit exits this  
condition is entered in this location. The  
Default value is -80.  
High Differential Pressure Alarm Setpoint  
The setpoint at which the unit determines that  
there is a High Differential Pressure alarm  
condition. The Default value is 1500.  
Low Differential Pressure Alarm Setpoint  
The setpoint at which the unit determines that  
there is a Low Differential Pressure alarm  
condition. The Default value is -100.  
High Differential Pressure Alarm Reset  
After a High Differential Pressure alarm  
occurs, the setpoint at which the unit exits this  
condition is entered in this location. The  
Default value is 1480.  
Low Flow Rate Alarm Reset  
After a Low Flow Rate alarm occurs, the  
setpoint at which the unit exits this condition  
is entered in this location. The Default value  
is -80.  
High Flow Rate Alarm Reset  
After a High Flow Rate alarm occurs, the  
setpoint at which the unit exits this condition  
is entered in this location. The Default value  
is 99,990.  
Low Flow Rate Alarm Setpoint  
The setpoint at which the unit determines that  
there is a Low Flow Rate alarm condition.  
The Default value is -100.  
High Flow Rate Alarm Setpoint  
The setpoint at which the unit determines that  
there is a High Flow Rate alarm condition.  
The Default value is 100,000.  
Low Pressure Alarm Reset  
After a Low Pressure alarm occurs, the  
setpoint at which the unit exits this condition  
is entered in this location. The Default value  
is -80.  
High Pressure Alarm Setpoint  
The setpoint at which the unit determines that  
there is a High Pressure alarm condition. The  
Default value is 1500.  
C - 4  
January 2002  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
Low Pressure Alarm Setpoint  
Percent N2  
The setpoint at which the unit determines that  
there is a Low Pressure alarm condition. The  
Default value is -100.  
This parameter reflects the content of nitrogen  
(N2) currently present in the gas. This  
number should be updated only after taking  
an analysis. If the content is unknown, a zero  
(0) should be entered. The Range is 0 to 0.15  
(15%). The Default is 0.  
Low Supply Voltage Alarm Reset  
After a Low Supply Voltage alarm occurs, the  
setpoint at which the unit exits this condition  
is entered in this location. The Default is 8.0.  
Percent O2  
This parameter reflects the content of oxygen  
(O2) currently present in the gas. This  
number should be updated only after taking  
an analysis. If the content is unknown, a zero  
(0) should be entered. The Range is 0 to 0.15  
(15%). The Default is 0.  
Low Supply Voltage Alarm Setpoint  
The setpoint at which the unit determines that  
there is a Low Supply Voltage alarm  
condition. The Default is 8.5.  
Low Temperature Alarm Reset  
After a Low Temperature alarm occurs, the  
setpoint at which the unit exits this condition  
is entered in this location. The Default is -  
80.  
Pressure Corrected Pulse Enable/Disable  
Corrected (press only) pulses can be sent to  
an external device by enabling the Pulse  
Output Channel. The Press determines the  
rate at which pulses are generated. Corr. Pulse  
Output - CF per Pulse parameter. To enable  
Relay #3 for pulse output, enter a value of  
35.3. An optional relay is required for the  
pulse output. Enter -1 to disable. The  
Default value is -1.  
Low Temperature Alarm Setpoint  
The setpoint at which the unit determines that  
there is a Low Temperature alarm condition.  
The Default is -100.  
Meter Correction Factor  
Press. Corr. Pulse Output - CF per Pulse  
This parameter should reflect the value of the  
pressure (only) corrected pulse output sent  
from the station to an external device. For  
example, if each output pulse represents 1000  
cubic feet, then this parameter should be set  
to 1000. The Default value is 1000.  
The Meter Correction Factor parameter will  
ordinarily be 1. The actual number is a ratio  
that indicates the measurement accuracy of  
the meter to which the device is attached. A  
setting of 1 indicates that the meter exhibits  
no measurement error. This parameter should  
not be changed from the default value of 1,  
unless the meter has been tested and its exact  
measurement error is known. The range is  
0.95 to 1.05. The Default value is 1.  
Site I.D. (RUID)  
The Site I.D. is unique to each device. It is  
the access code number that allows the user  
with a portable or Host computer to  
communicate with the unit. The Range is 1 to  
65,535. The Factory Default value is 1.  
Percent CO2  
This parameter reflects the content of carbon  
dioxide (CO2) currently present in the gas.  
This number should be updated only after  
taking an analysis. If the content is unknown,  
a zero (0) should be entered. The Range is 0  
to 0.15 (15%). The Default is 0.  
Specific Gravity  
Enter the specific gravity at the station. The  
Range is 0.554 and 2.000 inclusive. The  
Default value is 0.6.  
January 2002  
C - 5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
Supercompressibilty Calculated or Fixed  
This parameter is used to set the mode for  
supercompressibility calculations. If it is set  
to 0, then a new supercompressibility value  
will be calculated each time the process  
executes. If set to 1, the unit will use the  
value set in the Fixed Supercompressibility  
Uncorrected Volume & Flow Multiplier  
This multiplier specifies the output value for  
uncorrected volume and flow. For example,  
if the device is to measure flow in thousands  
of cubic feet (MCF), select Thousands  
(1000's) of units from the pick list. The  
Default value is 1000.  
Value parameter for calculations.  
The  
Default value is 0. (Calculated)  
Ten Thousands (10,000's) of units 10000  
Thousands (1000's) of units  
Hundreds (100's) of units  
Tens (10's) of units  
1000  
100  
10  
Fixed Supercompressibility Value  
If supercompressibility is Fixed, this  
parameter should be set to the desired fixed  
value. If supercompressibility is Calculated,  
any value entered will be ignored. The  
Default is 1.  
Single (1's) units  
1
Wake Up Interval (seconds)  
This parameter specifies the time (in seconds)  
that the unit will wake up, execute the station  
processes and perform all calculations.  
Because the setting of this parameter directly  
affects battery life, care should be exercised  
to ensure that this item is set to the longest  
interval possible for satisfactory operation.  
The default values are as follow:  
Time HHMMSS (Hours, Minutes,  
Seconds)  
This is the current military time in the unit. It  
is updated on each process scan.  
Uncorrected Pulse Out Enable/Disable  
Uncorrected pulses can be sent to an external  
device by enabling the Pulse Output Channel.  
The Uncorrected Pulse Output - CF per Pulse  
parameter, determines the rate at which pulses  
are generated. An optional relay is required  
for the pulse output. To enable Relay #4 for  
pulse output, enter a value of 35.2. Enter -1  
to disable. The Default value is -1. (Disable)  
Battery or Solar unit  
600 seconds - (Use if the unit powers up and  
down and the Wake up on pulse parameter  
(EFC only) is set to 0). The unit will wake up  
every 10 minutes to execute the station  
processes and calculate a new flow rate.  
3600 seconds - (Use if the Wake up on pulse  
parameter is not 0 to allow for history  
recording - EFC’s only).  
Uncorr Pulse Output - CF per Pulse  
This parameter should reflect the value of the  
uncorrected pulse output from the station to  
an external device. For example, if each  
output pulse represents 1000 cubic feet, then  
this parameter should be set to 1000. The  
Default is 1000.  
AC or Solar units  
0 seconds - (Use if the unit is to be powered  
up at all times).  
performed on  
(approximately every 3.5 seconds).  
Calculations will be  
each process scan  
Wake Up On Pulse - event driven (EFC)  
If this parameter is greater than 0, the EFC  
will wake-up on the specified number of  
pulses (meter revolutions). During wake-up,  
pressure, temperature, and the rest of the  
analog channels are sampled, and the unit  
executes the processes and run calculations.  
Uncorrected Volume  
The uncorrected index represents the total  
index volume registered by the station. When  
changing this number, you will normally enter  
the number that appears on the mechanical  
index attached to the station. The Default  
value is 0.  
C - 6  
January 2002  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
It is disabled if set to 0. In this mode, the  
EFC should be configured to wake up on the  
number of pulses entered along with an  
hourly scheduled wake-up to record history  
data.  
Therefore, the wake up interval  
(seconds) parameter should be set to 3600.  
The Default value is 0.  
Units of Measure  
The MTEK6000 can be configured to calculate volume and flow with English or Metric units.  
The database ca be setup for the following units of measure:  
Flow & Volume  
Accumulations  
ft3 (cubic feet)  
M3 (cubic meters)  
Static Pressure  
PSI (pounds per square inch)  
Kpa (Kilopascals)  
Kg / cm2 (kilograms per square centimeter)  
Bars  
MBars (millibars)  
Atmospheres  
mmHg (millimeters of mercury)  
mmH2O (millimeters of water)  
MH2O (meters of water)  
inH2O (inches of water)  
Temperature*  
°F  
°C  
Differential  
Pressure  
inH2O (inches of water)  
mmH2O (mm of water)  
Pipe & Orifice  
Diameter  
inches  
centimeters  
* The units for temperature are automatically set for °F or °C when the units for volume and  
flow are chosen. Choosing English units (cubic feet) will set the temperature to F, etc.  
January 2002  
C - 7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
C - 8  
January 2002  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
APPENDIX D : BOARD JUMPER POSITIONS  
61-SBC Revision A – Corrector Board  
January 2002  
D-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
61-OPT Revision A – Option Board  
Pulse Output 1:  
Pulse Output 2:  
Use  
and  
Terminal 19 (normally open contact)  
Terminal 18 (common 1)  
Terminal 17 (normally closed contact)  
For form A output: put JP1 and JP1A both in position A to B.  
For form C output: put JP1 and JP1A both in position B to C (requires external wetting).  
Use  
and  
Terminal 16 (normally open contact)  
Terminal 15 (common 2)  
Terminal 14 (normally closed contact)  
For form A output: put JP2 and JP2A both in position A to B.  
For form C output: put JP2 and JP2A both in position B to C (requires external wetting).  
Pulse Output 3 :  
(optional)  
Use  
Terminal 13 (normally open contact)  
Terminal 12 (common 3)  
and  
Terminal 11 (normally closed contact)  
For form A output: put JP3 and JP3A both in position A to B.  
For form C output: put JP3 and JP3A both in position B to C (requires external wetting).  
Pulse Output 4:  
(optional)  
Use  
and  
Terminal 10 (normally open contact)  
Terminal 9 (common 4)  
Terminal 8 (normally closed contact)  
For form A output: put JP4 and JP4A both in position A to B  
For form C output: put JP4 and JP4A both in position B to C (requires external wetting).  
JP5 can be used as a convenient means to connect pulse input 2 as follows:  
Description  
connect pulse input 2 to mechanical  
index secondary output  
otherwise  
Jumper position  
JP5 A, B, & C all in  
JP 5 A, B, & C all out (default)  
D-2  
January 2002  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
JP6 selects a power source for external analog transducers at TB3-1 & TB6-1 as follows:  
Description  
Jumper position  
JP6 = A  
use on board +12V dc supply ( 20 mA max )  
use on-board +5V dc supply ( 20 mA max)  
Vin external supply (from Metretek UPS)  
JP6 = B  
JP6 = C  
External Analog #2: Use  
Terminal 35 (transducer power)  
Terminal 36 (transducer output)  
Terminal 37 (transducer common)  
For a 4-to-20 mA type transducer put JP7 and JP8 in.  
For a 1-to-5 volt type transducer take JP7 and JP8 out.  
External Analog #1: Use  
Terminal 20 (transducer power)  
Terminal 21 (transducer output)  
Terminal 22 (transducer common)  
For a 4-to-20 mA type transducer put JP9 and JP10 in.  
For a 1-to-5 volt type transducer take JP9 and JP10 out.  
Pulse Input #1:  
Pulse Input #2:  
Use  
Terminal 32 (set input)  
Terminal 33 (ground)  
Terminal 34 (reset input)  
Use  
Terminal 29 (set input)  
Terminal 30 (ground)  
Terminal 31 (reset input)  
Status / Pulse Input #3: Use Terminal 26 (set input)  
Terminal 27 (ground)  
Terminal 28 (reset input)  
External Access to the Mechanical Index:  
Terminal 25 (normally open)  
(separate form C contacts)  
Terminal 24 (common)  
Terminal 23 (normally closed)  
January 2002  
D-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
50-PLI Revision C - Phone Line Interface  
J1: Connections from the PLI to the Corrector board  
J3: Telephone line connection  
J4: Earth Ground (for surge suppression) connection  
JP5: position 2-3 shorts R9 sometimes required when barriers cause too much drop in the OH  
signal level (default position is 1-2).  
D-4  
January 2002  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
many devices are attached, they may not ring  
properly.  
APPENDIX E : Certifications (CSA, UL  
and FCC Drawings and Statements  
CONSUMER INFORMATION AND FCC  
REQUIREMENTS  
Service Requirement  
5. In the event of equipment malfunction, all  
repairs should be performed by our  
Company or an authorized agent. It is the  
responsibility of users requiring services  
to report the need for service to our  
Company or to one of our authorized  
agents.  
1. The Federal Communication commission  
(FCC) has established rules, which  
permits this device to be directly  
connected to the telephone network.  
Standardized jacks are used for these  
connections. This equipment should not  
be used on party lines or coin lines.  
Service can be obtained at:  
2. If this device is malfunctioning, it may  
also be causing harm to the telephone  
network; this device should be  
disconnected until the source of the  
problem can be determined and unit repair  
as been made. If this is not done, the  
telephone company may temporarily  
disconnect service.  
Metretek, Inc.  
300 North Drive,  
Melbourne, Florida 32934  
Telephone: (321)-259-9700  
3. The telephone company may make  
changes in its technical operations and  
procedures. If such changes affect the  
compatibility or use of this device, the  
telephone company is required to give  
adequate notice of the changes.  
This device complies with Part 15 and Part 68  
of the FCC Rules. Operation is subjected to  
the following two conditions:  
[1] This device may not cause harmful  
Interference, and  
4. If the telephone company requests  
information on what equipment is  
connected to its lines, inform them of:  
[2] This device must accept any interference  
received, including interference that may  
cause undesired operations.  
(a) The telephone number that this unit is  
connected to  
(b) The ringer equivalence number  
[0.8B]  
(c) The USOC jack required [ RJ11C ]  
(d) The FCC Registration Number  
BK5 USA-35754-DT-T  
Items (b) and (d) are indicated on the label.  
The ringer equivalence number (REN) is used  
to determine how many devices can be  
connected to your telephone line. In most  
areas, the sum of the RENs of all devices on  
any line should not exceed five (5.0). If too  
January 2002  
E-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
CANADIAN “INDUSTRY CANADA”  
NOTICE  
NOTICE  
The Ringer Equivalence Number (REN)  
assigned to each terminal device provides an  
indication of the maximum number of  
terminals allowed to be connected to a  
telephone interface. The termination on an  
interface may consist of any combination of  
devices subject only to the requirement that  
the sum of the Ringer Equivalence Numbers  
of all the devices does not exceed 5.  
The Industry Canada label identifies certified  
equipment. This certification means that the  
equipment  
meets  
telecommunications  
network Protective, operational and safety  
requirements as prescribed in the appropriate  
Terminal Equipment Technical Requirements  
document(s).  
The Department does not  
guarantee the equipment will operate to the  
user’s satisfaction.  
Ringer Equivalence Number [0.8B]  
Before installing this equipment, users should  
ensure that it is permissible to be connected to  
the facilities of the local telecommunications  
company. The equipment must also be  
installed using an acceptable method of  
connection. The customer should be aware  
that compliance with the above conditions  
may not prevent degradation of service in  
some situations.  
WARRANTY & REPAIR SERVICE  
Repair & Return Department  
Metretek, Inc.  
300 North Drive  
Melbourne, Florida 32934  
(800) 327-8559  
Repairs to certified equipment should be  
coordinated by a representative designated by  
the supplier. Any repairs or alterations made  
by the user to this equipment, or equipment  
malfunctions,  
may  
give  
the  
telecommunications company cause to  
request the user to disconnect the equipment.  
Users should ensure for their own protection  
that the electrical ground connections of the  
power utility, telephone lines and internal  
metallic water pipe system, if present, are  
connected together. This precaution may be  
particularly important in rural areas.  
CAUTION  
Users should not attempt to make such  
connections themselves, but should contact  
the appropriate electric inspection authority,  
or an electrician, as appropriate.  
To be supplied when final Class 1, Div 1 & Div 2 approvals for this unit are received  
E-2  
January 2002  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
EL0001 (Sheet 1 of 2) : Class I Division 2 Installation  
January 2002  
E-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
EL0001 (Sheet 2 of 2) : Class I Division 2 Installation  
E-4  
January 2002  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
The seller makes no representation or  
warranty other than those set forth in this  
agreement. THE WARRANTY STATED  
HEREIN IS EXPRESSLY IN LIEU OF  
ALL WARRANTIES, EXPRESSED OR  
IMPLIED, INCLUDING BUT NOT  
LIMITED TO, ANY EXPRESSED OR  
APPENDIX F: Warranty Information  
WARRANTY INFORMATION  
The seller warrants its hardware to be free  
from defects in material and workmanship  
under normal and proper use for a period of  
one year from the date the hardware is  
shipped from Metretek, Incorporated. The  
seller’s sole liability and the buyer’s sole  
remedy for any breach of the foregoing  
provision is, at the seller’s option, the  
timely no-charge repair or replacement of  
any defective hardware or part that  
Metretek inspects and finds reasonable  
evidence that a defect in material or  
IMPLIED  
MERCHANTABILITY OR FITNESS  
FOR PARTICULAR PURPOSE.  
WARRANTY  
OF  
A
SUCH WARRANTY CONSTITUTES  
THE ONLY WARRANTY MADE BY  
THE SELLER WITH RESPECT TO  
THIS AGREEMENT, THE EQUIPMENT  
UNITS, OR THE SERVICES TO BE  
SUPPLIED HEREBY. THE SELLER  
SHALL NOT BE LIABLE FOR ANY  
INCIDENTAL OR CONSEQUENTIAL  
DAMAGES OF ANY KIND.  
workmanship exists.  
The buyer shall  
provide the labor required to remove the  
defective hardware and install its  
replacement at no charge to the seller. The  
equipment will be shipped to the seller at  
the buyer’s expense. The replacement or  
repaired equipment will be shipped to the  
buyer at the seller’s expense.  
The warranty will not extend to equipment  
subjected to accident, to misuse, or to  
alterations/repair  
not  
made  
and  
documented in writing by Metretek.  
Warranty claims to be honored under this  
warranty must be made promptly. Such  
claims shall specify the nature and details  
of the claim, the date that the cause of the  
claim was first observed, and the affected  
equipment’s unit serial number. Defective  
equipment shall not be returned to the  
seller’s factory without prior authorization  
from the seller. A copy of the claim’s  
documentation must be attached to the  
defective equipment and sent to the seller’s  
manufacturing  
facility.  
Defective  
components replaced under this warranty  
shall become the property of the seller.  
January 2002  
F-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
F-2  
January 2002  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
APPENDIX G: TII Station Protector  
January 2002  
G-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
G-2  
January 2002  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
APPENDIX H: Hazardous Area Installation Control Drawings  
October 2002  
H-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
October 2002  
H-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MTEK6000 SERIES USER'S MANUAL  
October 2002  
H-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Compaq Server DL380 G2 User Manual
Creative Home Theater System A220 User Manual
Fujitsu Flat Panel Television PDS5003U User Manual
Fujitsu Laptop 6000 SS User Manual
Garmin Camcorder 10 Jan User Manual
GE Monogram Washer Dryer WBSR3140 User Manual
GE Washer Dryer GCG1500L User Manual
Gigabyte Network Card GN WB01GS User Manual
Gitzo Camcorder Accessories G528 User Manual
Gitzo Camcorder Accessories G1345 User Manual