NurseCall Relay Unit
F.01U.252.722 | V1 | 2011.12
en User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
NurseCall Relay Unit
Table of Contents | en
3
Table of Contents
Safety instructions
Importance of safety instructions
Disregarding safety rules
Environmental conditions
General safety instructions
Observation and information
Description
General description
Top view
Bottom view
Front view
Relay Unit rear view
Relay Plus Unit rear view
Detailed description
Loudspeaker
Display
Keyboard
RS-485 interface
Antenna
Installation
Unpacking
List of contents
Installation
Generalities
Installation on a piece of furniture
Wall installation
Installing the antenna
Connecting to the mains
Setting the jumpers on the communication board on the Relay Plus Unit
Connecting the RS-485
Connecting the RS-232 to the Relay Plus Unit
Setting the 100 Ohm termination jumper
Programming
Generalities
Using the keyboard
Exit and cancel entries
Key not allowed
Locking and unlocking the keyboard
Programming time-out
First use
List of original factory settings
Parameters set by the NurseCall Main Unit
Locating mode and identification number
Bosch Security Systems
User Manual
F.01U.252.722 | V1 | 2011.12
Download from Www.Somanuals.com. All Manuals Search And Download.
4
en | Table of Contents
NurseCall Relay Unit
Special settings
Displaying the firmware version
Resetting all the parameters and buffers
Assistance and fire priority
Assistance and fire non priority
Special texts in German
Standard texts in German
Printer/display setup on the Relay Plus Unit
Access to the printer/display setup menu
Selection "None"
Selection "Printer"
Selection "DISPLAY ZETTLER"
Selection "DISPLAY NORSK"
Operation
Adjusting the loudspeaker volume
Consulting the alarm or event buffer
Switching between alarm and event buffers indication
Display indications
The green button does not work
Disconnecting a Relay Unit
Maintenance
Checking the system
Monitoring the power supply
Monitoring the backup battery
Cleaning
Parts replacement
Disassembling the unit
Backup battery replacing
Disposal
Disassembly
Returning to the manufacturer
Materials
Battery
Appendix
Electrical specifications
Dimensions and weight
Environmental conditions
List of criteria
Connectors
LINE socket (unit bottom)
Power socket (unit bottom)
RS-232 socket (Relay Plus Unit rear)
RS-485 socket (unit rear)
F.01U.252.722 | V1 | 2011.12
User Manual
Bosch Security Systems
Download from Www.Somanuals.com. All Manuals Search And Download.
NurseCall Relay Unit
Safety instructions | en
5
1
Safety instructions
NOTICE!
The user and installer should read and understand this chapter before any intervention on the
NurseCall Relay Unit.
1.1
Importance of safety instructions
Each safety and protection instruction in this manual must be adhered to in order to avoid
personnel injuries, property damages or environmental pollution.
In a similar manner, the legal bylaws, the measures in prevention of accidents and for the
protection of the environment, as well as the recognized technical rules aiming at appropriate
and safe working conditions which as applied in the country and at the place of use of the
NurseCall Relay Unit must be adhered to.
1.2
1.3
Disregarding safety rules
Disregarding the safety rules, as well as existing legal and technical regulations, may lead to
accidents, to property damages or to environmental pollution.
Environmental conditions
NOTICE!
The NurseCall Relay Unit must not be located near a water tap or any other source of water.
The electrical safety of the unit is only guaranteed if the electrical installation is conform to
the national regulation and if this installation works properly.
The NurseCall Relay Unit may not be used in buildings prone to fire and explosion hazards.
NOTICE!
The NurseCall Relay Unit may not be used under exposure to the direct sunlight, to heat, to
dust or to an excessive humidity (only use the equipment in a clean environment).
Install the NurseCall Relay Unit in a dry place, away from any source of heat.
CAUTION!
Interferences
Avoid immediate proximity to other electric devices such as a television.
1.4
General safety instructions
DANGER!
Electrocution
During maintenance operations, when the NurseCall Relay Unit is powered and its casing is
removed, it may not be left unattended.
CAUTION!
The NurseCall Relay Unit may only be connected to the electrical sources as described in
Bosch Security Systems
User Manual
F.01U.252.722 | V1 | 2011.12
Download from Www.Somanuals.com. All Manuals Search And Download.
6
en | Safety instructions
NurseCall Relay Unit
CAUTION!
Maintenance and repairs may only be performed in conformance with the instructions and by
authorized technical personnel only.
The sole possession of the user manual does not allow the personnel to perform any kind of
repair on the NurseCall Relay Unit.
Take into account all the warnings and follow all the instructions displayed on the NurseCall
Relay Unit and those which are printed in the documentation.
Never try to use replacement pieces other than those authorized by the manufacturer of the
NurseCall Relay Unit.
CAUTION!
It is mandatory to use the products specified in the present user manual to clean the
NurseCall Relay Unit. If you plan to use another product, only do so after having obtained the
authorization of the manufacturer.
DANGER!
The NurseCall Relay Unit contains highly sensitive electronic components. It should be opened
only in an ESD protected environment with respect to the following precautions:
1. Discharge yourself from electrostatic loads by touching a grounded conductive surface
before opening the unit.
2. Avoid touching conductive parts inside the unit if not absolutely necessary.
CAUTION!
Never let any liquid enter the system. In case of liquid spill inside the NurseCall Relay Unit, act
immediately as follows:
1. Switch off the unit using the main switch under the casing.
2. Unplug the power supply adaptor.
3. Dry up the unit.
4. Clean the unit.
5. Check that the unit switches on correctly.
1.4.1
Observation and information
In case of defective operation or any other technical incident for which no remedy is
described in this manual, please contact immediately your local representative.
F.01U.252.722 | V1 | 2011.12
User Manual
Bosch Security Systems
Download from Www.Somanuals.com. All Manuals Search And Download.
NurseCall Relay Unit
Description | en
7
2
Description
2.1
General description
2.1.1
Top view
1
2
3
7
6
5 4
1. Loudspeaker.
2. Display.
3. Keyboard, under the cover.
4. LED Indicator
5. Yellow button
Used to view more details about the event or alarm currently displayed (date and time,
position, etc...).
6. Green button
Used to acknowledge an alarm locally, see Section 5.2.3 The green button does not work,
7. Red button with light
This button is not used. Pressing the button does not activate a function.
The light blinks red during an alarm.
Bosch Security Systems
User Manual
F.01U.252.722 | V1 | 2011.12
Download from Www.Somanuals.com. All Manuals Search And Download.
8
en | Description
NurseCall Relay Unit
2.1.2
Bottom view
1
2
3
3
4
6
5
1. Identification label.
2. Cable channels.
3. Wall mounting holes (distance between holes, 157 mm).
4. ON/OFF switch.
5. LINE socket, used for firmware update.
6. 10V AC socket.
F.01U.252.722 | V1 | 2011.12
User Manual
Bosch Security Systems
Download from Www.Somanuals.com. All Manuals Search And Download.
NurseCall Relay Unit
Description | en
9
2.1.3
Front view
1
1. LED Indicator
Status
LED
Standby mode (normal operation)
Backup battery low
Green (permanent)
Green (blinking)
Green (flashing)
Red (blinking)
Power supply disconnected
Help, assistance or fire
2.1.4
Relay Unit rear view
1
2
1. RS-485 connector.
2. Antenna connector
2.1.5
Relay Plus Unit rear view
1
2
3
1. RS-232 connector, used for the printer/display interface of the Relay Plus Unit.
2. RS-485 connector.
3. Antenna connector
Bosch Security Systems
User Manual
F.01U.252.722 | V1 | 2011.12
Download from Www.Somanuals.com. All Manuals Search And Download.
10 en | Description
NurseCall Relay Unit
2.2
Detailed description
2.2.1
Loudspeaker
When one of the following alarms or messages is received by the NurseCall Relay Unit, the
internal loudspeaker is activated until acknowledgement.
Status
Loudspeaker
Power supply disconnected
Call for help, reserve call, technical call
Error message
Dual-tone beep every minute
4 second interval, one tone
15 second interval, one tone
1 minute interval, one tone
Continuously dual-tone beep
Short beep
Disconnection of a relay unit from RS485-bus
Call for assistance / fire alarm
Local acknowledgement
2.2.2
Display
The NurseCall Relay Unit is equipped with a 2 x 20 characters display that guides the operator
during the programming. During normal operation, alarms and messages are displayed.
NurseCall Relay Unit
V2.02 BN111.242.00A
NOTICE!
This user manual is written for the unit language English USA. Certain displays may differ for
the unit language English UK.
2.2.3
Keyboard
The keyboard has 21 alphanumeric keys. They are used to perform special commands or
during normal operation.
OK
OFF
ON
P
T
C
F.01U.252.722 | V1 | 2011.12
User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Bosch Security Systems
NurseCall Relay Unit
Description | en
11
Keys
Function
Access to the printer/display setup (Relay Plus Unit only)
Increase the volume of the loudspeaker.
Decrease the volume of the loudspeaker.
Scroll down to the previous alarm/event.
Scroll up to the next alarm/event.
Confirm a value or a command.
Cancel an entry or a command.
Check the status of the backup battery.
Not used.
Enter a value, such as special codes.
to
Not used.
Not used.
Enter a value or launch the event/alarm display mode.
Lock and unlock the keyboard.
then
Bosch Security Systems
User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
F.01U.252.722 | V1 | 2011.12
12 en | Description
NurseCall Relay Unit
2.2.4
RS-485 interface
One NurseCall Main Unit and up to 32 NurseCall Relay Units can be connected by a RS-485
bus. The bus must be connected to pins 2 and 5 of the RS-485 socket.
NOTICE!
Keep polarity equal when connecting further units to the RS485 bus!
NOTICE!
Maximum RS485-bus length: 1200 m.
Use only one twisted pair cable for the interconnection.
NOTICE!
The receiver units located at the two ends of the bus must be terminated with a 100 Ohm
jumper setting.
In this configuration, you always should connect the NurseCall Main Unit first. The NurseCall
Relay Units must then be connected to the RS485-bus one by one, not at the same time.
Relay output
In the same connector, a potential free contact is available. It is a low current switching
contact. The relay (potential free, switching power max. 48 V / 0.5 A) is activated at a call for
help, call for assistance or fire alarm. This relay can be set as closing or switching contact
(cycle of 10 seconds on / 10 seconds off). This feature can be used to drive a signal lamp for
example.
For relay setting, see the user manual of the NurseCall Main Unit.
2.2.5
Antenna
The antenna is connected to the NurseCall Relay Unit using the adapter supplied with the
unit.
F.01U.252.722 | V1 | 2011.12
User Manual
Bosch Security Systems
Download from Www.Somanuals.com. All Manuals Search And Download.
NurseCall Relay Unit
Installation | en
13
3
Installation
3.1
Unpacking
The NurseCall Relay Unit is carefully packed for transportation.
The components contained in the box are protected, but should be handled with care.
Store the packaging material for further use (storage or transport).
1. Take all components out of the box and place the NurseCall Relay Unit on the working
space.
2. Check each component in the box, in accordance with the list of contents below.
3. Check that the NurseCall Relay Unit and its accessories have not been damaged during
transportation.
In case of defective or missing equipment, do not try to install the NurseCall Relay Unit.
Contact immediately your local representative.
3.1.1
List of contents
Description
NurseCall Relay Unit
Power supply adaptor (Europe)
230VAC/10VAC
or
Power supply adaptor (UK)
230VAC/10VAC UK
or
Power supply adaptor (US)
115VAC/10VAC
Antenna 434MHz 1/2 L=34 cm FME
Straight adapter BFME-TNC
Right angled bended adapter BFME-ETNC
2 m Cable FCC 6/4
NurseCall Relay Unit User Manual
Bosch Security Systems
User Manual
F.01U.252.722 | V1 | 2011.12
Download from Www.Somanuals.com. All Manuals Search And Download.
14 en | Installation
NurseCall Relay Unit
3.2
Installation
3.2.1
Generalities
Install the NurseCall Relay Unit in a dry place, away from any source of heat.
Tools required:
–
–
Torx T20 screwdriver.
Torx T10 screwdriver.
3.2.2
3.2.3
Installation on a piece of furniture
It is recommended to place the NurseCall Relay Unit on a non-slippery surface. However, do
not place anything (blanket or lace) on top of the unit.
Wall installation
You can fasten the NurseCall Relay Unit on a smooth wall surface using two screws. The
distance between holes is 157 mm.
The power line cord should be placed inside the cable channels on the bottom of the
NurseCall Relay Unit.
3.2.4
Installing the antenna
1. Use the straight adapter (4) for wall installation or the right angled bended adapter (3)
for installation on a piece of furniture.
2
4
2
3
2. Fasten the adapter (3) or (4) on the antenna connector (1).
1
3. Fasten the antenna (2) on the adapter.
F.01U.252.722 | V1 | 2011.12
User Manual
Bosch Security Systems
Download from Www.Somanuals.com. All Manuals Search And Download.
NurseCall Relay Unit
Installation | en
15
3.2.5
Connecting to the mains
The NurseCall Relay Unit is powered by an adaptor (230/10 VAC or 115/10 VAC).
CAUTION!
In case of a different supply, the equipment must fulfill isolation requirements according to
EN60950 standard (last edition).
1. Plug the power adaptor into a power outlet placed near the unit. It should be easily
accessible at any time.
1
2. Connect the cable to the socket labeled 10V AC (1), under the unit.
Bosch Security Systems
User Manual
F.01U.252.722 | V1 | 2011.12
Download from Www.Somanuals.com. All Manuals Search And Download.
16 en | Installation
NurseCall Relay Unit
3.2.6
Setting the jumpers on the communication board on the Relay Plus Unit
3. Set the jumpers as required in your configuration. By default the jumpers are set for
connection to a Display Zettler.
Setting the jumpers for a Display Zettler:
Setting the jumpers for a printer setup or a Display Norsk:
4. Assemble the communication board and the unit. This is basically the reverse of the
F.01U.252.722 | V1 | 2011.12
User Manual
Bosch Security Systems
Download from Www.Somanuals.com. All Manuals Search And Download.
NurseCall Relay Unit
Installation | en
17
3.2.7
Connecting the RS-485
One NurseCall Main Unit and up to 32 NurseCall Relay Units can be connected to an RS-485
bus. Please contact a specialist for a correct installation.
CAUTION!
Do not use a star connection for the RS-485 network!
Incorrect connection:
Relay Unit 1
Main Unit
Relay Unit 2
Relay Unit n
Correct connection:
RS-485 bus (1 twisted pair cable)
100 Ω
100 Ω
Relay Unit n
Main Unit
Relay Unit 1
Relay Unit 2
NOTICE!
The NurseCall Main or Relay Units located at the two ends of the bus must be terminated with
a 100 Ohm resistor.
Bosch Security Systems
User Manual
F.01U.252.722 | V1 | 2011.12
Download from Www.Somanuals.com. All Manuals Search And Download.
18 en | Installation
NurseCall Relay Unit
3.2.8
Connecting the RS-232 to the Relay Plus Unit
Connect the printer or display device to the 9-pole SUB-D connector (1) at the rear part
of the housing of the Relay Plus Unit.
1
3.2.9
Setting the 100 Ohm termination jumper
Within the NurseCall Main or Relay Units, the RS-485 interface can be configured with a
jumper.
3. Put the 100 Ohm termination jumper J112 (1).
1
4. Assemble the communication board and the unit. This is basically the reverse of the
NOTICE!
If you do not want to disassemble the NurseCall Relay Unit, you also can short-out pins 3 and
4 of the connector. This has the same effect as the jumper setting described above.
F.01U.252.722 | V1 | 2011.12
User Manual
Bosch Security Systems
Download from Www.Somanuals.com. All Manuals Search And Download.
NurseCall Relay Unit
Programming | en
19
4
Programming
4.1
Generalities
4.1.1
Using the keyboard
Open the cover carefully and use the keyboard.
ON
OK
OFF
C
To access the special settings programming, press
three times quickly.
4.1.2
4.1.3
Exit and cancel entries
Press
once or several times.
Key not allowed
If you have pressed a key by mistake during the programming, a high-pitched beep is
generated.
4.1.4
Locking and unlocking the keyboard
Press
then
within one second to lock or unlock the keyboard.
This function locks only the keyboard to prevent any false manipulation. The colored buttons
on the left hand side are still available.
When the keyboard is locked, a small lock appears at the bottom right of the display:
DATE: 03.01.12 ¨
TIME: 12:12:31 ½A
4.1.5
Programming time-out
Programming of the NurseCall Relay Unit terminates automatically if no entries are made
on the keyboard for more than one minute.
Bosch Security Systems
User Manual
F.01U.252.722 | V1 | 2011.12
Download from Www.Somanuals.com. All Manuals Search And Download.
20 en | Programming
NurseCall Relay Unit
4.2
First use
4.2.1
List of original factory settings
Parameter
Original Factory Setting
Page
* Locating mode
On
* Identification number
Assistance and fire priority
Special texts in German
Printer/display setup (Relay Plus Unit only)
Speaker volume
No value
No
No
None
Midrange
* Locating mode and identification number are set at first use. To change a value, reset the
4.2.2
4.2.3
Parameters set by the NurseCall Main Unit
The parameters listed below are programmed in the Main Unit and are applied to the Relay
Unit. See the NurseCall Main Unit user manual.
–
–
–
–
–
Display mode
Unit language
Date and time
Local acknowledgement and the access code
Output relay function and mode
Locating mode and identification number
1. Disconnect the NurseCall Relay Unit from the RS-485 bus.
2. Switch the NurseCall Relay Unit on with the switch located at the bottom of the unit.
Wait for the following display to appear:
LOCATING ?
( OFF / ON )
3. Enable (ON) or disable (OFF) the indication of the transmitter position (locating mode)
on the NurseCall Relay Unit display. It must be the same as in the NurseCall Main Unit.
Press
to disable or
to enable the locating mode.
4. The identification number of the NurseCall Relay Unit must now be attributed.
RELAY UNIT ID: ·
( + / - )
The identification number is attributed by the NurseCall Main Unit upon connecting the RS-
485 bus corresponds to the order of connection. It can also be set manually. The range for
this value is (A, B, C, ..., Y, Z, a, b, c, d, e, f) for a maximum of 32 Relay Units.
Automatic attribution by the NurseCall Main Unit
5.a. Press
or the Red button to erase any previous selection. The black box remains.
or the Green button to
5.b. Connect the NurseCall Relay Unit to the RS-485 bus. Press
enable the automatic attribution. The next free identification number is attributed by the Main
Unit and is displayed on the Relay Unit at the bottom right.
F.01U.252.722 | V1 | 2011.12
User Manual
Bosch Security Systems
Download from Www.Somanuals.com. All Manuals Search And Download.
NurseCall Relay Unit
Programming | en
21
Manual selection of the NurseCall Main Unit
5.a. Scroll with and to set manually the identification number. Confirm with
or the Green button. The Relay Unit signalizes that it is not connected to the RS-485 bus.
NurseCall
RELAY UNIT OFF !!!
5.b. Connect the NurseCall Relay Unit to the RS-485 bus. The identification number is now
displayed at the bottom right.
4.3
Special settings
After pressing
three times quickly, you can enter special codes.
Enter Code
4.3.1
Displaying the firmware version
1. Type the code 194155.
2. The version of the firmware will be displayed for a few seconds.
Software REV A
V2.02 BN111.242.00A
4.3.2
Resetting all the parameters and buffers
CAUTION!
Disconnect the RS-485 bus before performing a reset. Reconnect it when you are finished.
NOTICE!
The following procedure resets the NurseCall Relay Unit to its original factory settings. It
erases all the parameters, as well as the alarms and events buffers. This reset is mandatory if
you wish to change the locating mode or the identification number.
1. Type the code 194156.
2. When the unit resets, it plays a short melody and displays a temporary message.
Please Wait...
3. After a few seconds, the unit goes back to the first use display.
4.3.3
Assistance and fire priority
This command sets the assistance call and the fire alarm as priority calls.
This means that alarms of these types will be displayed first.
1. Type the code 123991. The unit plays a melody and a confirmation message is displayed.
ASSISTANCE & FIRE
PRIORITY
Bosch Security Systems
User Manual
F.01U.252.722 | V1 | 2011.12
Download from Www.Somanuals.com. All Manuals Search And Download.
22 en | Programming
NurseCall Relay Unit
4.3.4
Assistance and fire non priority
This command sets the assistance call and the fire alarm as non-priority calls. This means that
the last alarm is displayed, whatever its type. This is the default value.
1. Type the code 123992. The unit plays a melody and a confirmation message is displayed.
ASSISTANCE & FIRE
NONPRIORITY
4.3.5
4.3.6
Special texts in German
This command sets special texts in German. The displayed criteria are:
–
–
BAD/WC instead of TECHNIK
HILFE-2 instead of NOTRUF2
1. Type the code 123007. The unit plays a melody and a confirmation message is displayed.
MULTITONE TEXTE
BAD/WC + HILFE-2
Standard texts in German
This command sets standard texts in German. This is the default value.
1. Type the code 123008. The unit plays a melody and a confirmation message is displayed.
STANDARDTEXTE
TECHNIK + NOTRUF2
4.4
Printer/display setup on the Relay Plus Unit
4.4.1
Access to the printer/display setup menu
NOTICE!
The printer/display setup can only be configured on a NurseCall Relay Plus Unit connected to
a NurseCall system via the RS-485 bus.
1. Press
then
to access the programming menu.
¬ OK:Program
¤:Info
2. Press
to enter the RS-232 output menu or
to check the RS-232 output setup.
¬ Parameter Nr. 00
RS-232 Output
£
3. Press
to enter the printer/display setup. Press
to escape.
4. Scroll with
and to select one of the following:
–
–
–
–
None
Printer
DISPLAY ZETTLER
DISPLAY NORSK
F.01U.252.722 | V1 | 2011.12
User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Bosch Security Systems
NurseCall Relay Unit
Programming | en
23
4.4.2
Selection "None"
if you have selected None, press
to confirm.
¬ RS-232 Output
None
£
In this case, there will be no printer/display output. This is the default setting.
4.4.3
Selection "Printer"
if you have selected Printer, press
to confirm.
¬ RS-232 Output
Printer
¡
1. Set the printing of all the daily messages sent by transmitters.
¬ Printer
24h printed ?:YES ¡
2. Scroll with
and
to select YES or NO. Default value is YES.
3. Confirm with
.
4.4.4
Selection "DISPLAY ZETTLER"
if you have selected DISPLAY ZETTLER, press
to confirm.
¬ RS-232 Output
DISPLAY ZETTLER ¡
1. Set the rotational display of alarms.
¬ DISPLAY ZETTLER
Rota.of Alarms:YES ¡
2. Select YES or NO with
and
. Default value is YES. Confirm with
.
3. Set the display of the clock. The format is HH:MM.
¬ DISPLAY ZETTLER
Time displayed:YES ¡
4. Scroll with
and
to select YES or NO. Default value is YES.
5. Confirm with
.
6. Configure the display of the 10 identification positions . The outcome is visible at the top.
DISPLAY :CR ff rr b
CRITER. 2 + 1 SPACES
a. Select a field with
b. Confirm with
and
, change a value with
and
.
.
Bosch Security Systems
User Manual
F.01U.252.722 | V1 | 2011.12
Download from Www.Somanuals.com. All Manuals Search And Download.
24 en | Programming
NurseCall Relay Unit
c. Choose 0 to 9 characters for the criterion, followed by 0 to 9 spaces.
Default values are 2 characters + 1 space.
7. Repeat the operation for
a. the floor (0 to 3 characters and 0 to 9 spaces), by default 2 characters and 1 space
b. the room (0 to 3 characters and 0 to 9 spaces), by default 2 characters and 1 space
c. the bed (0 or 1 character), by default 1 character.
NOTICE!
Configuration of the display is possible only in display mode FL,RO,BE (floor, room, bed).
If the NurseCall system is in display mode NUMBER, there is no need of further programming.
4.4.5
Selection "DISPLAY NORSK"
NOTICE!
The display mode NUMBER ZZZZ (alphanumeric display) is not compatible with the Display
Norsk. Make sure to choose another display mode when connecting a Display Norsk to the
Relay Plus Unit. See the NurseCall Main Unit user manual for more information.
if you have selected DISPLAY NORSK, press
to confirm.
¬ RS-232 Output
DISPLAY NORSK
¢
1. Set the rotational display of alarms.
¬ DISPLAY NORSK
Rota.of Alarms:YES ¡
2. Select YES or NO with
and
. Default value is YES. Confirm with
.
3. Configure the display of the 6 identification positions. The outcome is visible at the top.
DISPLAY : ffrrrb
FL: 2 +RO: 3 +BE:1
–
Select a field with
Confirm with
and
, change a value with
and
.
–
–
.
Choose 0 to 9 characters for the criterion, followed by 0 to 9 spaces.
Default values are 2 characters and 1 space.
4. Repeat the operation for
–
–
–
the floor (0 to 3 characters), by default 2 characters
the room (0 to 3 characters), by default 3 characters
the bed (0 or 1 character), by default 1 character.
NOTICE!
Configuration of the display is possible only in display mode FL,RO,BE (floor, room, bed).
If the NurseCall system is in display mode NUMBER, there is no need of further programming.
F.01U.252.722 | V1 | 2011.12
User Manual
Bosch Security Systems
Download from Www.Somanuals.com. All Manuals Search And Download.
NurseCall Relay Unit
Operation | en
25
5
Operation
5.1
Adjusting the loudspeaker volume
When the NurseCall Relay Unit is in standby mode:
Press
Press
to increase the volume.
to decrease the volume.
5.2
Consulting the alarm or event buffer
The NurseCall Relay Unit uses an alarm buffer and an event buffer for display indication.
The following alarms and messages are stored into the alarm buffer:
–
–
–
–
–
–
–
–
call for help
call for assistance
reserve call (call for help 2)
technical call
fire alarm
battery low message
error message
disconnection of a NurseCall Relay Unit from the RS-485 bus
If alarms are repeated, only the least recent entry remains in the buffer. The call for assistance
replaces the call for help, the reserve call and the technical call in the alarm buffer.
In addition to all the alarms, all possible entries are stored in the event buffer.
The following messages are directly stored into the event buffer:
–
–
–
–
–
–
–
–
–
–
–
–
–
acknowledgement N46, sent by an N46 Wall Transmitter, S35 or S37 Transmitters
acknowledgement by S35 or S37 Transmitters
local acknowledgement, an acknowledgement at the NurseCall Main Unit or Relay Unit
daily message check
personnel arrival message (A, B, C and D)
personnel departure message
power outage of a receiver unit
return of power at a receiver unit
backup battery low of a receiver unit
interruption of the RS-232 connection interface between the NurseCall system and a PC
return of the RS-232 connection interface between the NurseCall system and a PC
connection of a NurseCall Relay Unit to the RS-485 bus
transmission of the event "door" by a RAC Wireless Contact.
NOTICE!
The alarm and event buffers have a capacity of 18 or 100 entries. See the maximum number of
events buffered, in the NurseCall Main Unit user manual.
The event buffer will normally be filled with the last 18 or 100 entries.
In the alarm buffer, only the active alarms are present.
Bosch Security Systems
User Manual
F.01U.252.722 | V1 | 2011.12
Download from Www.Somanuals.com. All Manuals Search And Download.
26 en | Operation
NurseCall Relay Unit
5.2.1
Switching between alarm and event buffers indication
The alarm buffer is indicated by default. If you are in the event buffer, the unit changes
automatically to the alarm buffer after 1 minute without activity.
If there are no entries in the alarm buffer, the display shows the actual date and time.
To switch from alarm to event buffer and vice versa, press
Scroll the alarms or the events with and
Scroll the alarms only with the Green button.
.
.
5.2.2
Display indications
With the Yellow button, you can switch between three available information blocks.
The following information is displayed when an alarm or a message arrives:
First information block
–
In case of a "floor/room/bed" display mode:
1. criterion of the alarm or message
2. alarm (A) or event (E) followed by its order in the buffer
3. identification of the transmitter location (floor/room/bed numbers)
–
In case of "single number" display mode:
1. criterion of the alarm or message
2. alarm (A) or event (E) followed by its order in the buffer
3. identification of the transmitter location (three digits or four signs)
Second information block
4. date of the event;
5. time of the event;
6. main unit (space) or relay Unit (A...f) identification number;
7. quality of the received radio signal.
This information is visible in all display modes.
F.01U.252.722 | V1 | 2011.12
User Manual
Bosch Security Systems
Download from Www.Somanuals.com. All Manuals Search And Download.
NurseCall Relay Unit
Operation | en
27
Third information block
8. position of the last passed beacon, visible in all display modes. If no beacon is registered
or if the alarm is sent outside of the range of a beacon, the POS 000 will be displayed.
In the alarm buffer, the total number of entries is indicated on top at the right. You can
immediately see how many alarms are active. In this example, there are a total of three alarms
in the alarm buffer. In the event buffer, the position of the event in the buffer is indicated:
LOCAL ACK.
E01
06:05:11 13:33:05 C2
Here, E01 corresponds to the latest entry in the event buffer.
Unit displaying the current date and time:
DATE: 03.01.11 ¨
TIME: 12:12:31
Unit displaying an active alarm:
HELP
A01
NU:001
Unit displaying an event in the buffer:
LOCAL ACK.
NU:001
E04
5.2.3
5.2.4
The green button does not work
Problem: you have tried unsuccessfully to acknowledge an alarm with the Green button.
Cause: the local acknowledgement is disabled.
Solution: activate the local acknowledgement function.
See the NurseCall Main Unit user manual.
Disconnecting a Relay Unit
If a Relay Unit stops communicating with the Main Unit, an alarm "Relay Off" is generated on
the Main Unit. This alarm can only be acknowledged on the Main Unit by pressing the Green
button, followed by the code 45. Thereafter, an event "No Relay" is generated.
As soon as the Relay Unit communicates again with the Main Unit, the Event "Relay On" is
generated. This operation is independent of the local acknowledgement setting.
Bosch Security Systems
User Manual
F.01U.252.722 | V1 | 2011.12
Download from Www.Somanuals.com. All Manuals Search And Download.
28 en | Maintenance
NurseCall Relay Unit
6
Maintenance
6.1
Checking the system
Check the correct function of your NurseCall system.
Perform periodically an alarm test.
6.2
Monitoring the power supply
In case of a power failure, the NurseCall Relay Unit emits a beep and the following message is
displayed alternatively with the date and time display:
Main Power
Error
The backup battery ensures that the NurseCall Relay Unit remains operational even in the case
of a power failure. When fully charged, the battery ensures a power backup of 24 hours.
When power returns after a power failure, the battery is recharged. If it has been completely
discharged, it will reach its full capacity after 24 hours of charging time.
6.3
Monitoring the backup battery
The status of the battery is indicated on the top right of the display.
To check the backup battery voltage, press
. The following message is displayed:
Checking
Local Battery
NOTICE!
At startup, an automatic check is made.
During normal operation, an automatic check is made every 30 minutes.
If the remaining battery capacity drops below 25 %, the following message is displayed:
Local Battery
Empty
If the NurseCall Relay Unit detects that the backup battery is defective, the following
message is displayed:
Local Battery
Failure
NOTICE!
If the backup battery is defective, replace it as described in Section 6.5.2 Backup battery
F.01U.252.722 | V1 | 2011.12
User Manual
Bosch Security Systems
Download from Www.Somanuals.com. All Manuals Search And Download.
NurseCall Relay Unit
Maintenance | en
29
6.4
Cleaning
Avoid using cleaning products or detergents.
Wipe off your NurseCall Relay Unit occasionally with a dry cloth.
6.5
Parts replacement
6.5.1
Disassembling the unit
Removing the antenna
1. Remove the antenna (1) and its adapter (2) or (3).
1
3
1
2
Removing the communication board
DANGER!
Do not damage the battery cable, its connector (7) or the serial communication board
connectors.
1. With a Torx T20 screwdriver, unscrew and remove the 4 screws (4).
2. With a Torx T10 screwdriver, unscrew and remove the screw (5).
3. Carefully remove the communication board (6).
4
6
5
7
Bosch Security Systems
User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
F.01U.252.722 | V1 | 2011.12
30 en | Maintenance
NurseCall Relay Unit
6.5.2
Backup battery replacing
Important Safety Instructions
The battery should charge for 24 hours before using the NurseCall Relay Unit for the first time,
after replacing the battery or after a long power shortage. Battery type is 6V NiMH.
NOTICE!
The battery will charge correctly between 5 ºC (41ºF) and 45 ºC (113 ºF). A battery that is
new or that has not been used for a long time can have reduced capacity at first use.
A rechargeable battery can be charged and discharged many times. However, it will eventually
wear out. This is not a defect. It is recommended to replace batteries that cannot ensure a
minimum power back-up time of 8 hours at full charge.
CAUTION!
–
–
–
–
–
–
–
–
–
–
May explode if exposed to fire.
Use only original batteries intended for your NurseCall Relay Unit.
Do not expose the battery to liquids.
Do not let the battery’s contacts touch another metal. This could damage the battery;
Do not disassemble or modify the battery;
Do not expose the battery to extreme temperatures, and never above 60 ºC (140 ºF).
For maximum battery capacity, use the battery at room temperature;
Keep out of reach of children;
Use the battery for the intended purpose only;
Do not put the battery in the mouth. Battery electrolytes may be toxic if swallowed.
CAUTION!
There is a risk of explosion if battery is replaced by a wrong type.
The battery should be replaced exclusively by authorized personnel.
Dispose of used batteries according to instructions and regulations.
Procedure
2. Disconnect the battery cable (1).
3. Carefully remove the backup battery (2).
4. Place the new backup battery.
5. Connect the new battery cable (1).
2
1
F.01U.252.722 | V1 | 2011.12
User Manual
Bosch Security Systems
Download from Www.Somanuals.com. All Manuals Search And Download.
NurseCall Relay Unit
Disposal | en
31
7
Disposal
The NurseCall Relay Unit is marked with a crossed-out wastebasket symbol.
This means that, at the end of its lifetime, the product should be disposed separately from
ordinary household waste in accordance with the EU Directive 2002/96/EC.
The product and its accessories should be delivered to an appropriate collection facility that
ensure recycling, treatment and an environmentally compatible disposal. This prevents any
negative impact on the environment and human health, and promotes the recycling of
materials. For more information on available collection facilities, contact your local waste
collection service or your local representative.
7.1
7.2
Disassembly
Only authorized personnel are allowed to disassemble a NurseCall Relay Unit.
Returning to the manufacturer
If there is no practical disposal place, the NurseCall Relay Unit may be returned to your local
representative.
7.3
7.4
Materials
The NurseCall Relay Unit must be returned to an authorized point of recycling.
In order to protect people and environment, the NurseCall Relay Unit must be recycled in an
adequate manner. Consequently, all applicable laws and bylaws must be respected.
Battery
NOTICE!
The battery should be disposed of as household waste. Use a battery disposal facility when
available.
Please check local regulations for disposal of batteries or call your local representative
for information.
Bosch Security Systems
User Manual
F.01U.252.722 | V1 | 2011.12
Download from Www.Somanuals.com. All Manuals Search And Download.
32 en |
NurseCall Relay Unit
A
Appendix
A.1
Electrical specifications
Voltage
Current
Frequency
Power
230 or 115/10VAC
280 mA
50/60 Hz
2.8 W max.
A.2
Dimensions and weight
Casing dimensions
Depth
[mm]
220
180
40
Width
Height
Antenna
[mm]
Height
400
Casing weight
[g]
Weight (including antenna and power supply adaptor)
740
A.3
Environmental conditions
Operating temperature
0 - 40°C
F.01U.252.722 | V1 | 2011.12
User Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Bosch Security Systems
NurseCall Relay Unit
| en
33
A.4
List of criteria
Criterion
Number
00
01
02
03
04
05
06
07
08
09
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
xx
Alarm (A) or Event (E) Comment
ERROR
A
E
A
E
E
A
E
A
E
E
E
A
E
A
E
E
A
E
E
E
E
E
E
E
E
A
E
E
E
E
E
E
E
E
A
System malfunction, e.g. component defective
Coded key active (N46)
Battery at low level (Transmitter)
Acknowledgement (Sent by N46, S35 or S37)
Door open or door closed
Technical Call (N46)
PERSONNEL A
LOW BATTERY
ACK. N46
DOOR
TECHNICAL
PERSONNEL C
HELP
Coded key presence (N46)
Call for Help
UNKNOWN
Not used
END PERSONNEL
PERSONNEL D
ASSISTANCE
24 HOURS
Removed coded key (N46)
Coded key presence (N46)
Assistance Call
Daily message check
RESERVE
Reserve Call (N46)
PERSONNEL B
RADIO NOISE
FIRE
Coded key presence (N46)
Bad radio transmission (noise)
Fire Alarm
ACK. TRANSM.1
LOCAL ACK.
POWER OUTAGE
POWER BACK
COMPUTER OFF
COMPUTER ON
BAT.ACK.TRANSM
RELAY ON
Acknowledgement (Ack. Transmitter No. 1)
Local Acknowledgement (Main Unit or Relay Unit)
Main Unit or Relay Unit not powered
Power back (Main Unit or Relay Unit)
Alarm management PC off
Alarm management PC on
Battery at low level (One of the Ack. Transmitter)
Relay Unit connected on RS-485 Bus
Relay Unit disconnected from RS-485 Bus
Relay Unit off acknowledged (Main Unit)
Accumulator discharged (Main Unit or Relay Unit)
Acknowledgement (Ack. Transmitter No. 2)
Acknowledgement (Ack. Transmitter No. 3)
Acknowledgement (Ack. Transmitter No. 4)
Acknowledgement (Ack. Transmitter No. 5)
Acknowledgement (Ack. Transmitter No. xx)
Acknowledgement (Ack. Transmitter No. 32)
Dementia Alarm
RELAY OFF
NO RELAY
LOW ACCU
ACK. TRANSM.2
ACK. TRANSM.3
ACK. TRANSM.4
ACK. TRANSM.5
ACK. TRANSM.xx
ACK. TRANSM.32
DEMENTIA
58
60
NOTICE!
All events are buffered into the event buffer of the NurseCall Relay Unit. All alarms and events
are sent to the printer/display. All alarms and events except the events "COMPUTER OFF" and
"COMPUTER ON" are sent to the Alarm management Software.
Bosch Security Systems
User Manual
F.01U.252.722 | V1 | 2011.12
Download from Www.Somanuals.com. All Manuals Search And Download.
34 en |
NurseCall Relay Unit
A.5
Connectors
A.5.1
LINE socket (unit bottom)
LINE socket
Wiring
1. Flash Data GND
2. Not used
3. Not used
4. Not used
5. Not used
6. Flash Data IN/OUT
1 2 3 4 5 6
A.5.2
Power socket (unit bottom)
10V AC socket
Wiring
1. Not used
2. AC-1 10-12VAC
3. AC-2
4. GND
1 2 3 4
A.5.3
RS-232 socket (Relay Plus Unit rear)
RS-232 socket
Wiring
1. ---
5 4 3 2 1
2. TXD (RXD)
3. RXD (TXD)
4. ---
9 8 7 6
5. GND
6. ---
7. ---
8. ---
9. ---
A.5.4
RS-485 socket (unit rear)
RS-485 socket
Wiring
1. Relay output (a)
2. RS485 (A)
1
2
3
4
5
6
3. Termination = RS485 (A)
4. RS485(A) when jumper end line is placed
5. RS485 (B)
6. Relay Output (b)
F.01U.252.722 | V1 | 2011.12
User Manual