Bosch Appliances Home Care Product F01U252722 User Manual

NurseCall Relay Unit  
F.01U.252.722 | V1 | 2011.12  
en User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NurseCall Relay Unit  
Table of Contents | en  
3
Table of Contents  
Safety instructions  
Importance of safety instructions  
Disregarding safety rules  
Environmental conditions  
General safety instructions  
Observation and information  
Description  
General description  
Top view  
Bottom view  
Front view  
Relay Unit rear view  
Relay Plus Unit rear view  
Detailed description  
Loudspeaker  
Display  
Keyboard  
RS-485 interface  
Antenna  
Installation  
Unpacking  
List of contents  
Installation  
Generalities  
Installation on a piece of furniture  
Wall installation  
Installing the antenna  
Connecting to the mains  
Setting the jumpers on the communication board on the Relay Plus Unit  
Connecting the RS-485  
Connecting the RS-232 to the Relay Plus Unit  
Setting the 100 Ohm termination jumper  
Programming  
Generalities  
Using the keyboard  
Exit and cancel entries  
Key not allowed  
Locking and unlocking the keyboard  
Programming time-out  
First use  
List of original factory settings  
Parameters set by the NurseCall Main Unit  
Locating mode and identification number  
Bosch Security Systems  
User Manual  
F.01U.252.722 | V1 | 2011.12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4
en | Table of Contents  
NurseCall Relay Unit  
Special settings  
Displaying the firmware version  
Resetting all the parameters and buffers  
Assistance and fire priority  
Assistance and fire non priority  
Special texts in German  
Standard texts in German  
Printer/display setup on the Relay Plus Unit  
Access to the printer/display setup menu  
Selection "None"  
Selection "Printer"  
Selection "DISPLAY ZETTLER"  
Selection "DISPLAY NORSK"  
Operation  
Adjusting the loudspeaker volume  
Consulting the alarm or event buffer  
Switching between alarm and event buffers indication  
Display indications  
The green button does not work  
Disconnecting a Relay Unit  
Maintenance  
Checking the system  
Monitoring the power supply  
Monitoring the backup battery  
Cleaning  
Parts replacement  
Disassembling the unit  
Backup battery replacing  
Disposal  
Disassembly  
Returning to the manufacturer  
Materials  
Battery  
Appendix  
Electrical specifications  
Dimensions and weight  
Environmental conditions  
List of criteria  
Connectors  
LINE socket (unit bottom)  
Power socket (unit bottom)  
RS-232 socket (Relay Plus Unit rear)  
RS-485 socket (unit rear)  
F.01U.252.722 | V1 | 2011.12  
User Manual  
Bosch Security Systems  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NurseCall Relay Unit  
Safety instructions | en  
5
1
Safety instructions  
NOTICE!  
The user and installer should read and understand this chapter before any intervention on the  
NurseCall Relay Unit.  
1.1  
Importance of safety instructions  
Each safety and protection instruction in this manual must be adhered to in order to avoid  
personnel injuries, property damages or environmental pollution.  
In a similar manner, the legal bylaws, the measures in prevention of accidents and for the  
protection of the environment, as well as the recognized technical rules aiming at appropriate  
and safe working conditions which as applied in the country and at the place of use of the  
NurseCall Relay Unit must be adhered to.  
1.2  
1.3  
Disregarding safety rules  
Disregarding the safety rules, as well as existing legal and technical regulations, may lead to  
accidents, to property damages or to environmental pollution.  
Environmental conditions  
NOTICE!  
The NurseCall Relay Unit must not be located near a water tap or any other source of water.  
The electrical safety of the unit is only guaranteed if the electrical installation is conform to  
the national regulation and if this installation works properly.  
The NurseCall Relay Unit may not be used in buildings prone to fire and explosion hazards.  
NOTICE!  
The NurseCall Relay Unit may not be used under exposure to the direct sunlight, to heat, to  
dust or to an excessive humidity (only use the equipment in a clean environment).  
Install the NurseCall Relay Unit in a dry place, away from any source of heat.  
CAUTION!  
Interferences  
Avoid immediate proximity to other electric devices such as a television.  
1.4  
General safety instructions  
DANGER!  
Electrocution  
During maintenance operations, when the NurseCall Relay Unit is powered and its casing is  
removed, it may not be left unattended.  
CAUTION!  
The NurseCall Relay Unit may only be connected to the electrical sources as described in  
Bosch Security Systems  
User Manual  
F.01U.252.722 | V1 | 2011.12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
6
en | Safety instructions  
NurseCall Relay Unit  
CAUTION!  
Maintenance and repairs may only be performed in conformance with the instructions and by  
authorized technical personnel only.  
The sole possession of the user manual does not allow the personnel to perform any kind of  
repair on the NurseCall Relay Unit.  
Take into account all the warnings and follow all the instructions displayed on the NurseCall  
Relay Unit and those which are printed in the documentation.  
Never try to use replacement pieces other than those authorized by the manufacturer of the  
NurseCall Relay Unit.  
CAUTION!  
It is mandatory to use the products specified in the present user manual to clean the  
NurseCall Relay Unit. If you plan to use another product, only do so after having obtained the  
authorization of the manufacturer.  
DANGER!  
The NurseCall Relay Unit contains highly sensitive electronic components. It should be opened  
only in an ESD protected environment with respect to the following precautions:  
1. Discharge yourself from electrostatic loads by touching a grounded conductive surface  
before opening the unit.  
2. Avoid touching conductive parts inside the unit if not absolutely necessary.  
CAUTION!  
Never let any liquid enter the system. In case of liquid spill inside the NurseCall Relay Unit, act  
immediately as follows:  
1. Switch off the unit using the main switch under the casing.  
2. Unplug the power supply adaptor.  
3. Dry up the unit.  
4. Clean the unit.  
5. Check that the unit switches on correctly.  
1.4.1  
Observation and information  
In case of defective operation or any other technical incident for which no remedy is  
described in this manual, please contact immediately your local representative.  
F.01U.252.722 | V1 | 2011.12  
User Manual  
Bosch Security Systems  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
NurseCall Relay Unit  
Description | en  
7
2
Description  
2.1  
General description  
2.1.1  
Top view  
1
2
3
7
6
5 4  
1. Loudspeaker.  
2. Display.  
3. Keyboard, under the cover.  
4. LED Indicator  
5. Yellow button  
Used to view more details about the event or alarm currently displayed (date and time,  
position, etc...).  
6. Green button  
Used to acknowledge an alarm locally, see Section 5.2.3 The green button does not work,  
7. Red button with light  
This button is not used. Pressing the button does not activate a function.  
The light blinks red during an alarm.  
Bosch Security Systems  
User Manual  
F.01U.252.722 | V1 | 2011.12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
8
en | Description  
NurseCall Relay Unit  
2.1.2  
Bottom view  
1
2
3
3
4
6
5
1. Identification label.  
2. Cable channels.  
3. Wall mounting holes (distance between holes, 157 mm).  
See Section 3.2.3 Wall installation, page 14 for a detailed description.  
4. ON/OFF switch.  
5. LINE socket, used for firmware update.  
6. 10V AC socket.  
F.01U.252.722 | V1 | 2011.12  
User Manual  
Bosch Security Systems  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
NurseCall Relay Unit  
Description | en  
9
2.1.3  
Front view  
1
1. LED Indicator  
Status  
LED  
Standby mode (normal operation)  
Backup battery low  
Green (permanent)  
Green (blinking)  
Green (flashing)  
Red (blinking)  
Power supply disconnected  
Help, assistance or fire  
2.1.4  
Relay Unit rear view  
1
2
1. RS-485 connector.  
2. Antenna connector  
2.1.5  
Relay Plus Unit rear view  
1
2
3
1. RS-232 connector, used for the printer/display interface of the Relay Plus Unit.  
2. RS-485 connector.  
3. Antenna connector  
Bosch Security Systems  
User Manual  
F.01U.252.722 | V1 | 2011.12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
10 en | Description  
NurseCall Relay Unit  
2.2  
Detailed description  
2.2.1  
Loudspeaker  
When one of the following alarms or messages is received by the NurseCall Relay Unit, the  
internal loudspeaker is activated until acknowledgement.  
Status  
Loudspeaker  
Power supply disconnected  
Call for help, reserve call, technical call  
Error message  
Dual-tone beep every minute  
4 second interval, one tone  
15 second interval, one tone  
1 minute interval, one tone  
Continuously dual-tone beep  
Short beep  
Disconnection of a relay unit from RS485-bus  
Call for assistance / fire alarm  
Local acknowledgement  
2.2.2  
Display  
The NurseCall Relay Unit is equipped with a 2 x 20 characters display that guides the operator  
during the programming. During normal operation, alarms and messages are displayed.  
NurseCall Relay Unit  
V2.02 BN111.242.00A  
NOTICE!  
This user manual is written for the unit language English USA. Certain displays may differ for  
the unit language English UK.  
2.2.3  
Keyboard  
The keyboard has 21 alphanumeric keys. They are used to perform special commands or  
during normal operation.  
OK  
OFF  
ON  
P
T
C
F.01U.252.722 | V1 | 2011.12  
User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Bosch Security Systems  
             
NurseCall Relay Unit  
Description | en  
11  
Keys  
Function  
Access to the printer/display setup (Relay Plus Unit only)  
Increase the volume of the loudspeaker.  
Decrease the volume of the loudspeaker.  
Scroll down to the previous alarm/event.  
Scroll up to the next alarm/event.  
Confirm a value or a command.  
Cancel an entry or a command.  
Check the status of the backup battery.  
Not used.  
Enter a value, such as special codes.  
to  
Not used.  
Not used.  
Enter a value or launch the event/alarm display mode.  
Lock and unlock the keyboard.  
then  
Bosch Security Systems  
User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F.01U.252.722 | V1 | 2011.12  
12 en | Description  
NurseCall Relay Unit  
2.2.4  
RS-485 interface  
One NurseCall Main Unit and up to 32 NurseCall Relay Units can be connected by a RS-485  
bus. The bus must be connected to pins 2 and 5 of the RS-485 socket.  
NOTICE!  
Keep polarity equal when connecting further units to the RS485 bus!  
NOTICE!  
Maximum RS485-bus length: 1200 m.  
Use only one twisted pair cable for the interconnection.  
NOTICE!  
The receiver units located at the two ends of the bus must be terminated with a 100 Ohm  
resistor. See Section 3.2.7 Connecting the RS-485, page 17 for more information about the  
jumper setting.  
In this configuration, you always should connect the NurseCall Main Unit first. The NurseCall  
Relay Units must then be connected to the RS485-bus one by one, not at the same time.  
Relay output  
In the same connector, a potential free contact is available. It is a low current switching  
contact. The relay (potential free, switching power max. 48 V / 0.5 A) is activated at a call for  
help, call for assistance or fire alarm. This relay can be set as closing or switching contact  
(cycle of 10 seconds on / 10 seconds off). This feature can be used to drive a signal lamp for  
example.  
For relay setting, see the user manual of the NurseCall Main Unit.  
2.2.5  
Antenna  
The antenna is connected to the NurseCall Relay Unit using the adapter supplied with the  
unit.  
F.01U.252.722 | V1 | 2011.12  
User Manual  
Bosch Security Systems  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
NurseCall Relay Unit  
Installation | en  
13  
3
Installation  
3.1  
Unpacking  
The NurseCall Relay Unit is carefully packed for transportation.  
The components contained in the box are protected, but should be handled with care.  
Store the packaging material for further use (storage or transport).  
1. Take all components out of the box and place the NurseCall Relay Unit on the working  
space.  
2. Check each component in the box, in accordance with the list of contents below.  
3. Check that the NurseCall Relay Unit and its accessories have not been damaged during  
transportation.  
In case of defective or missing equipment, do not try to install the NurseCall Relay Unit.  
Contact immediately your local representative.  
3.1.1  
List of contents  
Description  
NurseCall Relay Unit  
Power supply adaptor (Europe)  
230VAC/10VAC  
or  
Power supply adaptor (UK)  
230VAC/10VAC UK  
or  
Power supply adaptor (US)  
115VAC/10VAC  
Antenna 434MHz 1/2 L=34 cm FME  
Straight adapter BFME-TNC  
Right angled bended adapter BFME-ETNC  
2 m Cable FCC 6/4  
NurseCall Relay Unit User Manual  
Bosch Security Systems  
User Manual  
F.01U.252.722 | V1 | 2011.12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
14 en | Installation  
NurseCall Relay Unit  
3.2  
Installation  
3.2.1  
Generalities  
Install the NurseCall Relay Unit in a dry place, away from any source of heat.  
Tools required:  
Torx T20 screwdriver.  
Torx T10 screwdriver.  
3.2.2  
3.2.3  
Installation on a piece of furniture  
It is recommended to place the NurseCall Relay Unit on a non-slippery surface. However, do  
not place anything (blanket or lace) on top of the unit.  
Wall installation  
You can fasten the NurseCall Relay Unit on a smooth wall surface using two screws. The  
distance between holes is 157 mm.  
The power line cord should be placed inside the cable channels on the bottom of the  
NurseCall Relay Unit.  
3.2.4  
Installing the antenna  
1. Use the straight adapter (4) for wall installation or the right angled bended adapter (3)  
for installation on a piece of furniture.  
2
4
2
3
2. Fasten the adapter (3) or (4) on the antenna connector (1).  
1
3. Fasten the antenna (2) on the adapter.  
F.01U.252.722 | V1 | 2011.12  
User Manual  
Bosch Security Systems  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
NurseCall Relay Unit  
Installation | en  
15  
3.2.5  
Connecting to the mains  
The NurseCall Relay Unit is powered by an adaptor (230/10 VAC or 115/10 VAC).  
CAUTION!  
In case of a different supply, the equipment must fulfill isolation requirements according to  
EN60950 standard (last edition).  
1. Plug the power adaptor into a power outlet placed near the unit. It should be easily  
accessible at any time.  
1
2. Connect the cable to the socket labeled 10V AC (1), under the unit.  
Bosch Security Systems  
User Manual  
F.01U.252.722 | V1 | 2011.12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
16 en | Installation  
NurseCall Relay Unit  
3.2.6  
Setting the jumpers on the communication board on the Relay Plus Unit  
1. Disassemble the unit; see Section 6.5.1 Disassembling the unit, page 29.  
2. Remove the communication board; see Section Removing the communication board,  
3. Set the jumpers as required in your configuration. By default the jumpers are set for  
connection to a Display Zettler.  
Setting the jumpers for a Display Zettler:  
Setting the jumpers for a printer setup or a Display Norsk:  
4. Assemble the communication board and the unit. This is basically the reverse of the  
disassembling procedure, see Section 6.5.1 Disassembling the unit, page 29.  
F.01U.252.722 | V1 | 2011.12  
User Manual  
Bosch Security Systems  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
NurseCall Relay Unit  
Installation | en  
17  
3.2.7  
Connecting the RS-485  
One NurseCall Main Unit and up to 32 NurseCall Relay Units can be connected to an RS-485  
bus. Please contact a specialist for a correct installation.  
CAUTION!  
Do not use a star connection for the RS-485 network!  
Incorrect connection:  
Relay Unit 1  
Main Unit  
Relay Unit 2  
Relay Unit n  
Correct connection:  
RS-485 bus (1 twisted pair cable)  
100 Ω  
100 Ω  
Relay Unit n  
Main Unit  
Relay Unit 1  
Relay Unit 2  
NOTICE!  
The NurseCall Main or Relay Units located at the two ends of the bus must be terminated with  
a 100 Ohm resistor.  
Bosch Security Systems  
User Manual  
F.01U.252.722 | V1 | 2011.12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
18 en | Installation  
NurseCall Relay Unit  
3.2.8  
Connecting the RS-232 to the Relay Plus Unit  
Connect the printer or display device to the 9-pole SUB-D connector (1) at the rear part  
of the housing of the Relay Plus Unit.  
1
3.2.9  
Setting the 100 Ohm termination jumper  
Within the NurseCall Main or Relay Units, the RS-485 interface can be configured with a  
jumper.  
1. Disassemble the unit as described in Section 6.5.1 Disassembling the unit, page 29.  
2. Remove the communication board as described in Section Removing the communication  
3. Put the 100 Ohm termination jumper J112 (1).  
1
4. Assemble the communication board and the unit. This is basically the reverse of the  
disassembling procedure, see Section 6.5.1 Disassembling the unit, page 29.  
NOTICE!  
If you do not want to disassemble the NurseCall Relay Unit, you also can short-out pins 3 and  
4 of the connector. This has the same effect as the jumper setting described above.  
F.01U.252.722 | V1 | 2011.12  
User Manual  
Bosch Security Systems  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
NurseCall Relay Unit  
Programming | en  
19  
4
Programming  
4.1  
Generalities  
4.1.1  
Using the keyboard  
Open the cover carefully and use the keyboard.  
ON  
OK  
OFF  
C
To access the special settings programming, press  
three times quickly.  
4.1.2  
4.1.3  
Exit and cancel entries  
Press  
once or several times.  
Key not allowed  
If you have pressed a key by mistake during the programming, a high-pitched beep is  
generated.  
4.1.4  
Locking and unlocking the keyboard  
Press  
then  
within one second to lock or unlock the keyboard.  
This function locks only the keyboard to prevent any false manipulation. The colored buttons  
on the left hand side are still available.  
When the keyboard is locked, a small lock appears at the bottom right of the display:  
DATE: 03.01.12 ¨  
TIME: 12:12:31 ½A  
4.1.5  
Programming time-out  
Programming of the NurseCall Relay Unit terminates automatically if no entries are made  
on the keyboard for more than one minute.  
Bosch Security Systems  
User Manual  
F.01U.252.722 | V1 | 2011.12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
             
20 en | Programming  
NurseCall Relay Unit  
4.2  
First use  
4.2.1  
List of original factory settings  
Parameter  
Original Factory Setting  
Page  
* Locating mode  
On  
* Identification number  
Assistance and fire priority  
Special texts in German  
Printer/display setup (Relay Plus Unit only)  
Speaker volume  
No value  
No  
No  
None  
Midrange  
* Locating mode and identification number are set at first use. To change a value, reset the  
4.2.2  
4.2.3  
Parameters set by the NurseCall Main Unit  
The parameters listed below are programmed in the Main Unit and are applied to the Relay  
Unit. See the NurseCall Main Unit user manual.  
Display mode  
Unit language  
Date and time  
Local acknowledgement and the access code  
Output relay function and mode  
Locating mode and identification number  
1. Disconnect the NurseCall Relay Unit from the RS-485 bus.  
2. Switch the NurseCall Relay Unit on with the switch located at the bottom of the unit.  
Wait for the following display to appear:  
LOCATING ?  
( OFF / ON )  
3. Enable (ON) or disable (OFF) the indication of the transmitter position (locating mode)  
on the NurseCall Relay Unit display. It must be the same as in the NurseCall Main Unit.  
Press  
to disable or  
to enable the locating mode.  
4. The identification number of the NurseCall Relay Unit must now be attributed.  
RELAY UNIT ID: ·  
( + / - )  
The identification number is attributed by the NurseCall Main Unit upon connecting the RS-  
485 bus corresponds to the order of connection. It can also be set manually. The range for  
this value is (A, B, C, ..., Y, Z, a, b, c, d, e, f) for a maximum of 32 Relay Units.  
Automatic attribution by the NurseCall Main Unit  
5.a. Press  
or the Red button to erase any previous selection. The black box remains.  
or the Green button to  
5.b. Connect the NurseCall Relay Unit to the RS-485 bus. Press  
enable the automatic attribution. The next free identification number is attributed by the Main  
Unit and is displayed on the Relay Unit at the bottom right.  
F.01U.252.722 | V1 | 2011.12  
User Manual  
Bosch Security Systems  
Download from Www.Somanuals.com. All Manuals Search And Download.  
           
NurseCall Relay Unit  
Programming | en  
21  
Manual selection of the NurseCall Main Unit  
5.a. Scroll with and to set manually the identification number. Confirm with  
or the Green button. The Relay Unit signalizes that it is not connected to the RS-485 bus.  
NurseCall  
RELAY UNIT OFF !!!  
5.b. Connect the NurseCall Relay Unit to the RS-485 bus. The identification number is now  
displayed at the bottom right.  
4.3  
Special settings  
After pressing  
three times quickly, you can enter special codes.  
Enter Code  
4.3.1  
Displaying the firmware version  
1. Type the code 194155.  
2. The version of the firmware will be displayed for a few seconds.  
Software REV A  
V2.02 BN111.242.00A  
4.3.2  
Resetting all the parameters and buffers  
CAUTION!  
Disconnect the RS-485 bus before performing a reset. Reconnect it when you are finished.  
NOTICE!  
The following procedure resets the NurseCall Relay Unit to its original factory settings. It  
erases all the parameters, as well as the alarms and events buffers. This reset is mandatory if  
you wish to change the locating mode or the identification number.  
1. Type the code 194156.  
2. When the unit resets, it plays a short melody and displays a temporary message.  
Please Wait...  
3. After a few seconds, the unit goes back to the first use display.  
4. Select the locating mode and the identification number. See Section 4.2 First use.  
4.3.3  
Assistance and fire priority  
This command sets the assistance call and the fire alarm as priority calls.  
This means that alarms of these types will be displayed first.  
1. Type the code 123991. The unit plays a melody and a confirmation message is displayed.  
ASSISTANCE & FIRE  
PRIORITY  
Bosch Security Systems  
User Manual  
F.01U.252.722 | V1 | 2011.12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
             
22 en | Programming  
NurseCall Relay Unit  
4.3.4  
Assistance and fire non priority  
This command sets the assistance call and the fire alarm as non-priority calls. This means that  
the last alarm is displayed, whatever its type. This is the default value.  
1. Type the code 123992. The unit plays a melody and a confirmation message is displayed.  
ASSISTANCE & FIRE  
NONPRIORITY  
4.3.5  
4.3.6  
Special texts in German  
This command sets special texts in German. The displayed criteria are:  
BAD/WC instead of TECHNIK  
HILFE-2 instead of NOTRUF2  
1. Type the code 123007. The unit plays a melody and a confirmation message is displayed.  
MULTITONE TEXTE  
BAD/WC + HILFE-2  
Standard texts in German  
This command sets standard texts in German. This is the default value.  
1. Type the code 123008. The unit plays a melody and a confirmation message is displayed.  
STANDARDTEXTE  
TECHNIK + NOTRUF2  
4.4  
Printer/display setup on the Relay Plus Unit  
4.4.1  
Access to the printer/display setup menu  
NOTICE!  
The printer/display setup can only be configured on a NurseCall Relay Plus Unit connected to  
a NurseCall system via the RS-485 bus.  
1. Press  
then  
to access the programming menu.  
¬ OK:Program  
¤:Info  
2. Press  
to enter the RS-232 output menu or  
to check the RS-232 output setup.  
¬ Parameter Nr. 00  
RS-232 Output  
£
3. Press  
to enter the printer/display setup. Press  
to escape.  
4. Scroll with  
and to select one of the following:  
None  
Printer  
DISPLAY ZETTLER  
DISPLAY NORSK  
F.01U.252.722 | V1 | 2011.12  
User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Bosch Security Systems  
             
NurseCall Relay Unit  
Programming | en  
23  
4.4.2  
Selection "None"  
if you have selected None, press  
to confirm.  
¬ RS-232 Output  
None  
£
In this case, there will be no printer/display output. This is the default setting.  
4.4.3  
Selection "Printer"  
if you have selected Printer, press  
to confirm.  
¬ RS-232 Output  
Printer  
¡
1. Set the printing of all the daily messages sent by transmitters.  
¬ Printer  
24h printed ?:YES ¡  
2. Scroll with  
and  
to select YES or NO. Default value is YES.  
3. Confirm with  
.
4.4.4  
Selection "DISPLAY ZETTLER"  
if you have selected DISPLAY ZETTLER, press  
to confirm.  
¬ RS-232 Output  
DISPLAY ZETTLER ¡  
1. Set the rotational display of alarms.  
¬ DISPLAY ZETTLER  
Rota.of Alarms:YES ¡  
2. Select YES or NO with  
and  
. Default value is YES. Confirm with  
.
3. Set the display of the clock. The format is HH:MM.  
¬ DISPLAY ZETTLER  
Time displayed:YES ¡  
4. Scroll with  
and  
to select YES or NO. Default value is YES.  
5. Confirm with  
.
6. Configure the display of the 10 identification positions . The outcome is visible at the top.  
DISPLAY :CR ff rr b  
CRITER. 2 + 1 SPACES  
a. Select a field with  
b. Confirm with  
and  
, change a value with  
and  
.
.
Bosch Security Systems  
User Manual  
F.01U.252.722 | V1 | 2011.12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
24 en | Programming  
NurseCall Relay Unit  
c. Choose 0 to 9 characters for the criterion, followed by 0 to 9 spaces.  
Default values are 2 characters + 1 space.  
7. Repeat the operation for  
a. the floor (0 to 3 characters and 0 to 9 spaces), by default 2 characters and 1 space  
b. the room (0 to 3 characters and 0 to 9 spaces), by default 2 characters and 1 space  
c. the bed (0 or 1 character), by default 1 character.  
NOTICE!  
Configuration of the display is possible only in display mode FL,RO,BE (floor, room, bed).  
If the NurseCall system is in display mode NUMBER, there is no need of further programming.  
4.4.5  
Selection "DISPLAY NORSK"  
NOTICE!  
The display mode NUMBER ZZZZ (alphanumeric display) is not compatible with the Display  
Norsk. Make sure to choose another display mode when connecting a Display Norsk to the  
Relay Plus Unit. See the NurseCall Main Unit user manual for more information.  
if you have selected DISPLAY NORSK, press  
to confirm.  
¬ RS-232 Output  
DISPLAY NORSK  
¢
1. Set the rotational display of alarms.  
¬ DISPLAY NORSK  
Rota.of Alarms:YES ¡  
2. Select YES or NO with  
and  
. Default value is YES. Confirm with  
.
3. Configure the display of the 6 identification positions. The outcome is visible at the top.  
DISPLAY : ffrrrb  
FL: 2 +RO: 3 +BE:1  
Select a field with  
Confirm with  
and  
, change a value with  
and  
.
.
Choose 0 to 9 characters for the criterion, followed by 0 to 9 spaces.  
Default values are 2 characters and 1 space.  
4. Repeat the operation for  
the floor (0 to 3 characters), by default 2 characters  
the room (0 to 3 characters), by default 3 characters  
the bed (0 or 1 character), by default 1 character.  
NOTICE!  
Configuration of the display is possible only in display mode FL,RO,BE (floor, room, bed).  
If the NurseCall system is in display mode NUMBER, there is no need of further programming.  
F.01U.252.722 | V1 | 2011.12  
User Manual  
Bosch Security Systems  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
NurseCall Relay Unit  
Operation | en  
25  
5
Operation  
5.1  
Adjusting the loudspeaker volume  
When the NurseCall Relay Unit is in standby mode:  
Press  
Press  
to increase the volume.  
to decrease the volume.  
5.2  
Consulting the alarm or event buffer  
The NurseCall Relay Unit uses an alarm buffer and an event buffer for display indication.  
The following alarms and messages are stored into the alarm buffer:  
call for help  
call for assistance  
reserve call (call for help 2)  
technical call  
fire alarm  
battery low message  
error message  
disconnection of a NurseCall Relay Unit from the RS-485 bus  
If alarms are repeated, only the least recent entry remains in the buffer. The call for assistance  
replaces the call for help, the reserve call and the technical call in the alarm buffer.  
In addition to all the alarms, all possible entries are stored in the event buffer.  
The following messages are directly stored into the event buffer:  
acknowledgement N46, sent by an N46 Wall Transmitter, S35 or S37 Transmitters  
acknowledgement by S35 or S37 Transmitters  
local acknowledgement, an acknowledgement at the NurseCall Main Unit or Relay Unit  
daily message check  
personnel arrival message (A, B, C and D)  
personnel departure message  
power outage of a receiver unit  
return of power at a receiver unit  
backup battery low of a receiver unit  
interruption of the RS-232 connection interface between the NurseCall system and a PC  
return of the RS-232 connection interface between the NurseCall system and a PC  
connection of a NurseCall Relay Unit to the RS-485 bus  
transmission of the event "door" by a RAC Wireless Contact.  
NOTICE!  
The alarm and event buffers have a capacity of 18 or 100 entries. See the maximum number of  
events buffered, in the NurseCall Main Unit user manual.  
The event buffer will normally be filled with the last 18 or 100 entries.  
In the alarm buffer, only the active alarms are present.  
Bosch Security Systems  
User Manual  
F.01U.252.722 | V1 | 2011.12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
26 en | Operation  
NurseCall Relay Unit  
5.2.1  
Switching between alarm and event buffers indication  
The alarm buffer is indicated by default. If you are in the event buffer, the unit changes  
automatically to the alarm buffer after 1 minute without activity.  
If there are no entries in the alarm buffer, the display shows the actual date and time.  
To switch from alarm to event buffer and vice versa, press  
Scroll the alarms or the events with and  
Scroll the alarms only with the Green button.  
.
.
5.2.2  
Display indications  
With the Yellow button, you can switch between three available information blocks.  
The following information is displayed when an alarm or a message arrives:  
First information block  
In case of a "floor/room/bed" display mode:  
1. criterion of the alarm or message  
2. alarm (A) or event (E) followed by its order in the buffer  
3. identification of the transmitter location (floor/room/bed numbers)  
In case of "single number" display mode:  
1. criterion of the alarm or message  
2. alarm (A) or event (E) followed by its order in the buffer  
3. identification of the transmitter location (three digits or four signs)  
Second information block  
4. date of the event;  
5. time of the event;  
6. main unit (space) or relay Unit (A...f) identification number;  
7. quality of the received radio signal.  
This information is visible in all display modes.  
F.01U.252.722 | V1 | 2011.12  
User Manual  
Bosch Security Systems  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
NurseCall Relay Unit  
Operation | en  
27  
Third information block  
8. position of the last passed beacon, visible in all display modes. If no beacon is registered  
or if the alarm is sent outside of the range of a beacon, the POS 000 will be displayed.  
In the alarm buffer, the total number of entries is indicated on top at the right. You can  
immediately see how many alarms are active. In this example, there are a total of three alarms  
in the alarm buffer. In the event buffer, the position of the event in the buffer is indicated:  
LOCAL ACK.  
E01  
06:05:11 13:33:05 C2  
Here, E01 corresponds to the latest entry in the event buffer.  
Unit displaying the current date and time:  
DATE: 03.01.11 ¨  
TIME: 12:12:31  
Unit displaying an active alarm:  
HELP  
A01  
NU:001  
Unit displaying an event in the buffer:  
LOCAL ACK.  
NU:001  
E04  
5.2.3  
5.2.4  
The green button does not work  
Problem: you have tried unsuccessfully to acknowledge an alarm with the Green button.  
Cause: the local acknowledgement is disabled.  
Solution: activate the local acknowledgement function.  
See the NurseCall Main Unit user manual.  
Disconnecting a Relay Unit  
If a Relay Unit stops communicating with the Main Unit, an alarm "Relay Off" is generated on  
the Main Unit. This alarm can only be acknowledged on the Main Unit by pressing the Green  
button, followed by the code 45. Thereafter, an event "No Relay" is generated.  
As soon as the Relay Unit communicates again with the Main Unit, the Event "Relay On" is  
generated. This operation is independent of the local acknowledgement setting.  
Bosch Security Systems  
User Manual  
F.01U.252.722 | V1 | 2011.12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
28 en | Maintenance  
NurseCall Relay Unit  
6
Maintenance  
6.1  
Checking the system  
Check the correct function of your NurseCall system.  
Perform periodically an alarm test.  
6.2  
Monitoring the power supply  
In case of a power failure, the NurseCall Relay Unit emits a beep and the following message is  
displayed alternatively with the date and time display:  
Main Power  
Error  
The backup battery ensures that the NurseCall Relay Unit remains operational even in the case  
of a power failure. When fully charged, the battery ensures a power backup of 24 hours.  
When power returns after a power failure, the battery is recharged. If it has been completely  
discharged, it will reach its full capacity after 24 hours of charging time.  
6.3  
Monitoring the backup battery  
The status of the battery is indicated on the top right of the display.  
To check the backup battery voltage, press  
. The following message is displayed:  
Checking  
Local Battery  
NOTICE!  
At startup, an automatic check is made.  
During normal operation, an automatic check is made every 30 minutes.  
If the remaining battery capacity drops below 25 %, the following message is displayed:  
Local Battery  
Empty  
If the NurseCall Relay Unit detects that the backup battery is defective, the following  
message is displayed:  
Local Battery  
Failure  
NOTICE!  
If the backup battery is defective, replace it as described in Section 6.5.2 Backup battery  
F.01U.252.722 | V1 | 2011.12  
User Manual  
Bosch Security Systems  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
NurseCall Relay Unit  
Maintenance | en  
29  
6.4  
Cleaning  
Avoid using cleaning products or detergents.  
Wipe off your NurseCall Relay Unit occasionally with a dry cloth.  
6.5  
Parts replacement  
6.5.1  
Disassembling the unit  
Removing the antenna  
1. Remove the antenna (1) and its adapter (2) or (3).  
1
3
1
2
Removing the communication board  
DANGER!  
Do not damage the battery cable, its connector (7) or the serial communication board  
connectors.  
1. With a Torx T20 screwdriver, unscrew and remove the 4 screws (4).  
2. With a Torx T10 screwdriver, unscrew and remove the screw (5).  
3. Carefully remove the communication board (6).  
4
6
5
7
Bosch Security Systems  
User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F.01U.252.722 | V1 | 2011.12  
         
30 en | Maintenance  
NurseCall Relay Unit  
6.5.2  
Backup battery replacing  
Important Safety Instructions  
The battery should charge for 24 hours before using the NurseCall Relay Unit for the first time,  
after replacing the battery or after a long power shortage. Battery type is 6V NiMH.  
NOTICE!  
The battery will charge correctly between 5 ºC (41ºF) and 45 ºC (113 ºF). A battery that is  
new or that has not been used for a long time can have reduced capacity at first use.  
A rechargeable battery can be charged and discharged many times. However, it will eventually  
wear out. This is not a defect. It is recommended to replace batteries that cannot ensure a  
minimum power back-up time of 8 hours at full charge.  
CAUTION!  
May explode if exposed to fire.  
Use only original batteries intended for your NurseCall Relay Unit.  
Do not expose the battery to liquids.  
Do not let the battery’s contacts touch another metal. This could damage the battery;  
Do not disassemble or modify the battery;  
Do not expose the battery to extreme temperatures, and never above 60 ºC (140 ºF).  
For maximum battery capacity, use the battery at room temperature;  
Keep out of reach of children;  
Use the battery for the intended purpose only;  
Do not put the battery in the mouth. Battery electrolytes may be toxic if swallowed.  
CAUTION!  
There is a risk of explosion if battery is replaced by a wrong type.  
The battery should be replaced exclusively by authorized personnel.  
Dispose of used batteries according to instructions and regulations.  
Procedure  
1. Disassemble the unit as described in Section 6.5.1 Disassembling the unit, page 29.  
2. Disconnect the battery cable (1).  
3. Carefully remove the backup battery (2).  
4. Place the new backup battery.  
5. Connect the new battery cable (1).  
2
1
F.01U.252.722 | V1 | 2011.12  
User Manual  
Bosch Security Systems  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
NurseCall Relay Unit  
Disposal | en  
31  
7
Disposal  
The NurseCall Relay Unit is marked with a crossed-out wastebasket symbol.  
This means that, at the end of its lifetime, the product should be disposed separately from  
ordinary household waste in accordance with the EU Directive 2002/96/EC.  
The product and its accessories should be delivered to an appropriate collection facility that  
ensure recycling, treatment and an environmentally compatible disposal. This prevents any  
negative impact on the environment and human health, and promotes the recycling of  
materials. For more information on available collection facilities, contact your local waste  
collection service or your local representative.  
7.1  
7.2  
Disassembly  
Only authorized personnel are allowed to disassemble a NurseCall Relay Unit.  
Returning to the manufacturer  
If there is no practical disposal place, the NurseCall Relay Unit may be returned to your local  
representative.  
7.3  
7.4  
Materials  
The NurseCall Relay Unit must be returned to an authorized point of recycling.  
In order to protect people and environment, the NurseCall Relay Unit must be recycled in an  
adequate manner. Consequently, all applicable laws and bylaws must be respected.  
Battery  
NOTICE!  
The battery should be disposed of as household waste. Use a battery disposal facility when  
available.  
Please check local regulations for disposal of batteries or call your local representative  
for information.  
Bosch Security Systems  
User Manual  
F.01U.252.722 | V1 | 2011.12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
32 en |  
NurseCall Relay Unit  
A
Appendix  
A.1  
Electrical specifications  
Voltage  
Current  
Frequency  
Power  
230 or 115/10VAC  
280 mA  
50/60 Hz  
2.8 W max.  
A.2  
Dimensions and weight  
Casing dimensions  
Depth  
[mm]  
220  
180  
40  
Width  
Height  
Antenna  
[mm]  
Height  
400  
Casing weight  
[g]  
Weight (including antenna and power supply adaptor)  
740  
A.3  
Environmental conditions  
Operating temperature  
0 - 40°C  
F.01U.252.722 | V1 | 2011.12  
User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Bosch Security Systems  
       
NurseCall Relay Unit  
| en  
33  
A.4  
List of criteria  
Criterion  
Number  
00  
01  
02  
03  
04  
05  
06  
07  
08  
09  
10  
11  
12  
13  
14  
15  
16  
17  
18  
19  
20  
21  
22  
23  
24  
25  
26  
27  
28  
29  
30  
31  
xx  
Alarm (A) or Event (E) Comment  
ERROR  
A
E
A
E
E
A
E
A
E
E
E
A
E
A
E
E
A
E
E
E
E
E
E
E
E
A
E
E
E
E
E
E
E
E
A
System malfunction, e.g. component defective  
Coded key active (N46)  
Battery at low level (Transmitter)  
Acknowledgement (Sent by N46, S35 or S37)  
Door open or door closed  
Technical Call (N46)  
PERSONNEL A  
LOW BATTERY  
ACK. N46  
DOOR  
TECHNICAL  
PERSONNEL C  
HELP  
Coded key presence (N46)  
Call for Help  
UNKNOWN  
Not used  
END PERSONNEL  
PERSONNEL D  
ASSISTANCE  
24 HOURS  
Removed coded key (N46)  
Coded key presence (N46)  
Assistance Call  
Daily message check  
RESERVE  
Reserve Call (N46)  
PERSONNEL B  
RADIO NOISE  
FIRE  
Coded key presence (N46)  
Bad radio transmission (noise)  
Fire Alarm  
ACK. TRANSM.1  
LOCAL ACK.  
POWER OUTAGE  
POWER BACK  
COMPUTER OFF  
COMPUTER ON  
BAT.ACK.TRANSM  
RELAY ON  
Acknowledgement (Ack. Transmitter No. 1)  
Local Acknowledgement (Main Unit or Relay Unit)  
Main Unit or Relay Unit not powered  
Power back (Main Unit or Relay Unit)  
Alarm management PC off  
Alarm management PC on  
Battery at low level (One of the Ack. Transmitter)  
Relay Unit connected on RS-485 Bus  
Relay Unit disconnected from RS-485 Bus  
Relay Unit off acknowledged (Main Unit)  
Accumulator discharged (Main Unit or Relay Unit)  
Acknowledgement (Ack. Transmitter No. 2)  
Acknowledgement (Ack. Transmitter No. 3)  
Acknowledgement (Ack. Transmitter No. 4)  
Acknowledgement (Ack. Transmitter No. 5)  
Acknowledgement (Ack. Transmitter No. xx)  
Acknowledgement (Ack. Transmitter No. 32)  
Dementia Alarm  
RELAY OFF  
NO RELAY  
LOW ACCU  
ACK. TRANSM.2  
ACK. TRANSM.3  
ACK. TRANSM.4  
ACK. TRANSM.5  
ACK. TRANSM.xx  
ACK. TRANSM.32  
DEMENTIA  
58  
60  
NOTICE!  
All events are buffered into the event buffer of the NurseCall Relay Unit. All alarms and events  
are sent to the printer/display. All alarms and events except the events "COMPUTER OFF" and  
"COMPUTER ON" are sent to the Alarm management Software.  
Bosch Security Systems  
User Manual  
F.01U.252.722 | V1 | 2011.12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
34 en |  
NurseCall Relay Unit  
A.5  
Connectors  
A.5.1  
LINE socket (unit bottom)  
LINE socket  
Wiring  
1. Flash Data GND  
2. Not used  
3. Not used  
4. Not used  
5. Not used  
6. Flash Data IN/OUT  
1 2 3 4 5 6  
A.5.2  
Power socket (unit bottom)  
10V AC socket  
Wiring  
1. Not used  
2. AC-1 10-12VAC  
3. AC-2  
4. GND  
1 2 3 4  
A.5.3  
RS-232 socket (Relay Plus Unit rear)  
RS-232 socket  
Wiring  
1. ---  
5 4 3 2 1  
2. TXD (RXD)  
3. RXD (TXD)  
4. ---  
9 8 7 6  
5. GND  
6. ---  
7. ---  
8. ---  
9. ---  
A.5.4  
RS-485 socket (unit rear)  
RS-485 socket  
Wiring  
1. Relay output (a)  
2. RS485 (A)  
1
2
3
4
5
6
3. Termination = RS485 (A)  
4. RS485(A) when jumper end line is placed  
5. RS485 (B)  
6. Relay Output (b)  
F.01U.252.722 | V1 | 2011.12  
User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Bosch Security Systems  
         
Download from Www.Somanuals.com. All Manuals Search And Download.  
Bosch Security Systems  
Robert-Bosch-Ring 5  
85630 Grasbrunn  
Germany  
© Bosch Security Systems, 2011  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Belkin Network Card N750 User Manual
Black Box Network Router phone line sharing device User Manual
Blaupunkt Car Stereo System 7 644 820 310 User Manual
BMW Automobile 318I User Manual
Bookendz Laptop Docking Station BE 10332 User Manual
Breville Juicer 800JEXL User Manual
Casio Clock TQ 379 User Manual
Casio Electronic Keyboard MA 170 User Manual
Casio Watch MA1201 EB User Manual
C Crane Digital Camera 024 S User Manual