Agilent Technologies Work Light 85054 90049 User Manual

User s a n d Ser vice Gu id e  
Agilen t Tech n ologies 85054B  
50Typ e-N Ca libr a tion Kit  
This manual applies directly to 85054B calibration kits with serial number  
prefix 3101A. The calibration devices in this kit are individually serialized.  
Record the device serial numbers in the table provided in this manual (see  
Recording the Device Serial Numbers” in Chapter 1.)  
Ma n u fa ct u r in g P a r t Nu m b er : 85054-90049  
P r in te d in USA  
P r in t Da t e: J u n e 2002  
Supersedes: Januray 1997  
© Copyright 1993, 1997, 2002 Agilent Technologies, Inc. All rights reserved.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Con t en t s  
iii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Con t en t s  
iv  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1 Gen er a l In for m a tion  
85054B  
1-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
General Information  
Calibration Kit Overview  
Ca lib r a tion Kit Over view  
The Agilent 85054B type-N calibration kit is used to calibrate Agilent network analyzers  
up to 18 GHz for measurements of components with 50type-N connectors.  
The standards in this calibration kit allow you to perform simple 1- or 2-port and TRM  
(thru–reflect–match) calibrations.  
This manual describes the 85054B calibration kit and provides replacement part numbers,  
specifications, and procedures for using, maintaining, and troubleshooting the kit.  
Kit Con ten ts  
Th e 85054B ca lib r a tion k it con t a in s t h e follow in g:  
offset opens and shorts, lowband and sliding load terminations  
four type-N to 7 mm adapters  
two type-N to type-N adapters  
a type-N connector gage set  
a 3/4 inch, 135 N-cm (12 in-lb) torque wrench for use on the type-N connectors  
a spanner wrench  
a data disk that contains the calibration definitions of the devices in the kit for 8510  
systems and the 872x series  
a data disk that contains the calibration definitions of the devices in the kit for the PNA  
series  
Refer to Table 6-1 and Figure 6-1 for a complete list of kit contents and their associated  
part numbers.  
Offse t Op en s a n d Sh or t s  
The offset opens and shorts are built from parts that are machined to the current  
state-of-the-art in precision machining.  
The offset shorts inner conductors have a one-piece construction, common with the  
shorting plane. The construction provides for extremely repeatable connections.  
The offset opens have inner conductors that are supported by a strong, low-dielectric  
constant plastic to minimize compensation values.  
Both the opens and shorts are constructed so that the pin depth can be controlled very  
tightly, thereby minimizing phase errors. The lengths of the offsets in the opens and shorts  
are designed so that the difference in phase of their reflection coefficients is approximately  
180 degrees at all frequencies.  
1-2  
85054B  
Download from Www.Somanuals.com. All Manuals Search And Download.  
           
General Information  
Calibration Kit Overview  
Low b a n d Loa d s  
The lowband loads are metrology-grade, 50 ohm terminations which have been optimized  
for lowband performance up to 2 GHz. The rugged internal structure provides for highly  
repeatable connections. A distributed resistive element on sapphire provides excellent  
stability and return loss.  
Slid in g Loa d s  
The sliding loads in this kit are designed to provide excellent performance from 3 GHz to  
18 GHz. The inner and outer conductors of the airline portion are precision machined to  
state-of-the-art tolerances. Although the sliding load has exceptional return loss, its  
superior load stability qualifies it as a high-performance device.  
The sliding load was designed with the ability to extend the inner conductor for connection  
purposes and then pull it back to a preset pin depth. This feature is critical since it  
minimizes the possibility of damage during the connection, while maintaining a minimum  
pin depth to optimize performance.  
Ad a p t er s  
Like the other devices in the kit, the adapters are built to very tight tolerances to provide  
good broadband performance. The adapters utilize a dual-beaded connector structure to  
ensure stable, repeatable connections. The beads are designed to minimize return loss and  
are separated far enough so that interaction between the beads is minimized.  
Ca lib r a tion Defin ition s  
The calibration kit must be selected and the calibration definitions for the devices in the  
kit installed in the network analyzer prior to performing a calibration. Refer to your  
network analyzer users guide for instructions on selecting the calibration kit and  
performing a calibration.  
The calibration definitions can be:  
resident within the analyzer  
loaded from the provided disk  
entered from the front panel  
In sta lla tion of t h e Ca lib r a tion Defin ition s  
The calibration definitions for the kit may be permanently installed in the internal  
memory or hard disk of the network analyzer.  
If the calibration definitions for the kit are not permanently installed in the network  
analyzer, they must be manually entered. Refer to your network analyzer users guide for  
instructions.  
Equ ip m en t Requ ir ed bu t Not Su p p lied  
Connector cleaning supplies and various electrostatic discharge (ESD) protection devices  
are not supplied with the calibration kit but are required to ensure successful operation of  
the kit. Refer to Table 6-2 on page 6-4 for ordering information.  
85054B  
1-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
           
General Information  
Incoming Inspection  
In com in g In sp ection  
Refer to Kit Contents” on page 1-2 to verify a complete shipment. Use Table 1-1 on page  
1-5 to record the serial numbers of all serialized devices in your kit.  
Check for damage. The foam-lined storage case provides protection during shipping. If the  
case or any device appears damaged, or if the shipment is incomplete, refer to Contacting  
Agilent” on page 5-3. Agilent will arrange for repair or replacement of incomplete or  
damaged shipments without waiting for a settlement from the transportation company.  
1-4  
85054B  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
General Information  
Serial Numbers  
Ser ia l Nu m ber s  
A serial number is attached to this calibration kit. The first four digits followed by a letter  
comprise the serial number prefix; the last five digits are the suffix, unique to each  
calibration kit.  
Recor d in g th e Device Ser ia l Nu m ber s  
In addition to the kit serial number, the devices in the kit are individually serialized (serial  
numbers are labeled onto the body of each device). Record these serial numbers in  
Table 1-1. Recording the serial numbers will prevent confusing the devices in this kit with  
similar devices from other kits.  
Ta b le 1-1 Ser ia l Nu m be r R ecor d for th e 85054B  
Device  
Ser ia l Nu m b er  
Ca lib r a t ion k it  
Lowband load (m)  
Lowband load (f)  
Open (m)  
_______________________________  
_______________________________  
_______________________________  
_______________________________  
_______________________________  
_______________________________  
_______________________________  
_______________________________  
_______________________________  
_______________________________  
Open (f)  
Short (m)  
Short (f)  
Sliding load (f)  
Sliding load (m)  
Connector gage (f)  
Ga ges  
Gage Master (f)  
Connector gage (m)  
Gage Master (m)  
Ad a p t er s  
_______________________________  
_______________________________  
_______________________________  
Type-N (m) to Type-N (m)  
Type-N (f) to Type-N (f)  
Type-N (f) to 7 mm  
Type-N (m) to 7 mm  
_______________________________  
_______________________________  
_______________________________  
_______________________________  
85054B  
1-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
General Information  
Calibration Kits Documented in This Manual  
Ca libr a tion Kits Docu m en t ed in Th is Ma n u a l  
This manual applies to any 85054B calibration kit whose serial number prefix is listed on  
the title page. If your calibration kit has a different serial number prefix, refer to the  
Calibration Kit History” section below for information on how this manual applies.  
Ca libr a tion Kit Histor y  
This section describes calibration kits with serial number prefixes lower that the ones  
listed on the title page.  
85054B Kit s w it h Ser ia l P r efix 2906A  
These calibration kits did not have the calibration definitions disk to support the Agilent  
8510C network analyzer. The part numbers provided in this manual are the recommended  
replacement parts for these kits. The devices in these kits should meet the specifications  
published in this manual.  
1-6  
85054B  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
General Information  
Precision Slotless Connectors  
P r ecision Slot less Con n ect or s  
The female type-N connectors in this calibration kit are metrology-grade, precision slotless  
connectors (PSC). A characteristic of metrology-grade connectors is direct traceability to  
national measurement standards through their well-defined mechanical dimensions.  
Conventiona l female center conductors are slotted. When mated, the female center  
conductor is flared by the male pin. Because physical dimensions determine connector  
impedance, electrical characteristics of the female connector (and connection pair) are  
dependent upon the mechanical dimensions of the male pin. While connectors are used in  
pairs, their male and female halves are always specified separately as part of a standard,  
instrument, or device under test. Because of these facts, making precision measurements  
with the conventional slotted connector is very difficult, and establishing a direct  
traceability path to primary dimensional standards is nearly impossible.  
The precision slotless connector was developed to eliminate these problems. All PSCs are  
female. A PSC incorporates a center conductor with a solid cylindrical shell that defines  
the outside diameter of the female center pin. Its outside diameter and, therefore, the  
impedance in its region does not change. The inner part provides an internal contact that  
flexes to accept the allowed range of male pin diameters.  
The calibration of a network analyzer having a conventional slotted female connector on  
the test port remains valid only when the device under test and all calibration standards  
have identical male pin diameters. For this reason, PSC test-port adapters are supplied in  
most calibration kits.  
Precision slotless connectors have the following characteristics:  
there is no loss of traceable calibration on test ports when the male pin diameter of the  
connector on the device under test is different from the male pin diameter of the  
calibration standard.  
The female PSC and its mating male connector can be measured and specified  
separately as part of the device either is attached to.  
All female connectors can have a known, stable impedance based only on the diameters  
of their inner and outer conductors.  
Female calibration standards can be fully specified. Their specifications and traceability  
are unaffected by the diameter of the male mating pin.  
A fully traceable performance verification is made using a precision 50 ohm airline  
having a PSC.  
Measurement repeatability is enhanced due to non-changing connector characteristics  
with various pin diameters.  
With PSCs on test ports and standards, the percentage of accuracy achieved when  
measuring at 50 dB return loss levels is comparable to using conventional slotted  
connectors measuring devices having only 30 dB return loss. This represents an accuracy  
improvement of about 10 times.  
85054B  
1-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
General Information  
Clarifying the Sex of a Connector  
Cla r ifyin g th e Sex of a Con n ect or  
In this manual, the sex of calibration devices and adapters are referred to in terms of their  
connector interface. For example, a male open has a male connector.  
However, during a measurement calibration, the network analyzer softkey menus label a  
type-N calibration device with reference to the sex of the analyzers test port  
connector—not the calibration device connector. For example, the label SHORT(F) on the  
analyzers display refers to the short that is to be connected to the female test port. This  
will be a male short from the calibration kit.  
Conversely, connector gages are referred to in terms of the connector that it measures. For  
instance, a male connector gage has a female connector on the gage so that it can measure  
male devices.  
P r even t ive Ma in ten a n ce  
The best techniques for maintaining the integrity of the devices in this kit include:  
routine visual inspection  
cleaning  
proper gaging  
proper connection techniques  
Failure to detect and remove dirt or metallic particles on a mating plane surface can  
degrade repeatability and accuracy and can damage any connector mated to it. Improper  
connections, resulting from pin depth values being out of the observed limits (see Table 2-2  
on page 2-4), or from bad connections, can also damage these devices.  
1-8  
85054B  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
2 Sp ecifica tion s  
2-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Specifications  
Environmental Requirements  
En vir on m en ta l R eq u ir em en ts  
Ta b le 2-1 E n vir on m en ta l R eq u ir em en ts  
Parameter  
Limits  
a
+20 °C to +26 °C (+68 °F to +79 °F)  
±1 °C of measurement calibration temperature  
40 °C to +75 °C (40 °F to +167 °F)  
Operating temperature  
b
Error-corrected temperature range  
Storage temperature  
Altitude  
Operation  
< 4,500 meters (15,000 feet)  
< 15,000 meters (50,000 feet)  
Always non-condensing  
0 to 80% (26 °C maximum dry bulb)  
0 to 90%  
Storage  
Relative humidity  
Operation  
Storage  
a. The temperature range over which the calibration standards maintain conformance to their  
specifications.  
b. The allowable network analyzer ambient temperature drift during measurement calibration  
and during measurements when the network analyzer error correction is turned on. Also, the  
range over which the network analyzer maintains its specified performance while correction  
is turned on.  
Tem p er a tu r eWh a t to Wa tch Ou t For  
Changes in temperature can affect electrical characteristics. Therefore, the operating  
temperature is a critical factor in performance. During a measurement calibration, the  
temperature of the calibration devices must be stable and within the range specified in  
IMPORTANT Avoid unnecessary handling of the devices during calibration because your  
fingers are a heat source.  
2-2  
85054B  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Specifications  
Mechanical Characteristics  
Mech a n ica l Ch a r a ct er ist ics  
Mechanical characteristics such as center conductor protrusion and pin depth are not  
performance specifications. They are, however, important supplemental characteristics  
related to electrical performance. Agilent Technologies verifies the mechanical  
characteristics of the devices in this kit with special gaging processes and electrical  
testing. This ensures that the device connectors do not exhibit any improper pin depth  
when the kit leaves the factory.  
Gaging Connectors” on page 3-6 explains how to use gages to determine if the kit devices  
have maintained their mechanical integrity. (Refer to Table 2-2 on page 2-4 for typical and  
observed pin depth limits.)  
P in Dep th  
Pin depth is the distance the center conductor mating plane differs from being flush with  
the outer conductor mating plane. Refer to Figure 2-1. Some coaxial connectors, such as  
2.4 mm and 3.5 mm, are designed to have these planes nearly flush. Type-N connectors,  
however, are designed with a pin depth offset of approximately 5.26 mm (0.207 inch), not  
permitting these planes to be flush. The male center conductors are recessed by the offset  
value while the female center conductors compensate by protruding the same amount.  
This offset necessitates the redefining of pin depth with regard to protrusion and  
recession.  
P r otr u sion refers to a male type-N connector center conductor having a pin depth value  
less than 5.26 mm (0.207 inch), or a female type-N connector center conductor having a pin  
depth value greater than 5.26 mm (0.207 inch).  
R ece ssion refers to a male type-N connector center conductor having a pin depth value  
greater than 5.26 mm (0.207 in), or a female type-N connector center conductor having a  
pin depth value less than 5.26 mm (0.207 inch).  
F igu r e 2-1  
Con n ect or P in Dep th  
85054B  
2-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
               
Specifications  
Mechanical Characteristics  
NOTE  
The gages for measuring type-N connectors compensate for the designed  
offset of 5.26 mm (0.207 inch), therefore, protrusion and recession readings  
are in relation to a zero reference plane (as if the inner and outer conductor  
planes were intended to be flush). Gage readings can be directly compared  
with the observed values listed in Table 2-2.  
The pin depth value of each calibration device in this kit is not specified, but is an  
important mechanical parameter. The electrical performance of the device depends, to  
some extent, on its pin depth. The electrical specifications for each device in this kit take  
into account the effect of pin depth on the devices performance. Table 2-2 lists the typical  
pin depths and measurement uncertainties, and provides observed pin depth limits for the  
devices in the kit. If the pin depth of a device does not measure within the observed pin  
depth limits, it may be an indication that the device fails to meet electrical specifications.  
Refer to Figure 2-1 for an illustration of pin depth in type-N connectors.  
Ta b le 2-2 P in Dep th Lim its  
Measurement  
Observed Pin Depth  
Device  
Typical Pin Depth  
micrometers  
a
b
Uncertainty  
Limits  
–4  
micrometers  
micrometers  
(10 inches)  
–4  
–4  
(10 inches)  
(10 inches)  
Opens  
Shorts  
0 to 12.7  
(0 to 5.0)  
+3.8 to 3.8  
+3.8 to 16.5  
(+ 1.5 to 6.5)  
(+ 1.5 to 1.5)  
0 to 12.7  
(0 to 5.0)  
+3.8 to 3.8  
(+ 1.5 to 1.5)  
+3.8 to 16.5  
(+ 1.5 to 21.5)  
Lowband  
loads  
0 to 50.8  
(0 to 20.0)  
+3.8 to 3.8  
(+ 1.5 to 1.5)  
+3.8 to 54.6  
(+ 1.5 to 21.5)  
Sliding loads  
0 to 7.6  
(0 to 3.0)  
+3.8 to 3.8  
(+ 1.5 to 1.5)  
+3.8 to 11.4  
(+ 1.5 to 4.5)  
Adapters  
(7 mm end)  
0 to 50.8  
(0 to 20.0)  
+3.8 to 3.8  
(+ 1.5 to 1.5)  
+3.8 to 54.6  
(+ 1.5 to 21.5)  
Adapters  
(type-N end)  
0 to 12.7  
(0 to 5.0)  
+3.8 to 3.8  
(+ 1.5 to 1.5)  
+3.8 to 16.5  
(+ 1.5 to 6.5)  
a. Approximately +2 sigma to 2 sigma of gage uncertainty based on studies done at the fac-  
tory according to recommended procedures.  
b. Observed pin depth limits are the range of observation limits seen on the gage reading due  
to measurement uncertainty. The depth could still be within specifications.  
NOTE  
When measuring pin depth, the measured value (resultant average of three  
or more measurements) is not the true value. Always compare the measured  
value with the observed pin depth limits in Table 2-2 to evaluate the  
condition of device connectors.  
2-4  
85054B  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Specifications  
Electrical Specifications  
E lect r ica l Sp ecifica t ion s  
The electrical specifications in Table 2-3 apply to the devices in your calibration kit when  
connected with an Agilent precision interface.  
Ta b le 2-3 E le ct r ica l Sp ecifica tion s  
Device  
Frequency (GHz)  
DC to 2  
Parameter  
Specification  
Lowband loads  
Return Loss  
Return Loss  
48 dB (0.00398ρ)  
42 dB (0.00794ρ)  
a
> 2 to 18  
Sliding loads  
34 dB (0.0200ρ)  
28 dB (0.0398ρ)  
Adapters  
DC to 8  
> 8 to 18  
at 18  
Return Loss  
Return Loss  
(both styles)  
b
Deviation from Nominal  
Phase  
±1.5°  
±1.0°  
Offset Opens  
at 18  
Deviation from Nominal  
Phase  
Offset Shorts  
a. The specifications for the sliding load termination include the quality of the airline  
portions within the sliding load combined with the effective stability of the sliding  
element.  
b. The specifications for the opens and shorts are given as allowed deviation from the  
nominal model as defined in the standard definitions (see Nominal Standard Definitions”  
Cer t ifica tion  
Agilent Technologies certifies that this product met its published specifications at the time  
of shipment from the factory. Agilent further certifies that its calibration measurements  
are traceable to the United States National Institute of Standards and Technology (NIST)  
to the extent allowed by the institutes calibration facility, and to the calibration facilities  
of other International Standards Organization members. See How Agilent Verifies the  
Devices in This Kit” on page 4-2 for more information.  
85054B  
2-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
           
Specifications  
Electrical Specifications  
2-6  
85054B  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3 Use, Ma in ten a n ce, a n d Ca r e of th e  
Devices  
3-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Use, Maintenance, and Care of the Devices  
Electrostatic Discharge  
Electr ost a t ic Disch a r ge  
Protection against ESD (electrostatic discharge) is essential while connecting, inspecting,  
or cleaning connectors attached to a static-sensitive circuit (such as those found in test  
sets).  
Static electricity can build up on your body and can easily damage sensitive internal  
circuit elements when discharged. Static discharges too small to be felt can cause  
permanent damage. Devices such as calibration components and devices under test  
(DUTs), can also carry an electrostatic charge. To prevent damage to the test set,  
components, and devices:  
a lwa ys wear a grounded wrist strap having a 1 Mresistor in series with it when  
handling components and devices or when making connections to the test set.  
a lwa ys use a grounded, conductive table mat while making connections.  
a lwa ys wear a heel strap when working in an area with a conductive floor. If you are  
uncertain about the conductivity of your floor, wear a heel strap.  
a lwa ys ground yourself before you clean, inspect, or make a connection to a  
static-sensitive device or test port. You can, for example, grasp the grounded outer shell  
of the test port or cable connector briefly.  
a lwa ys ground the center conductor of a test cable before making a connection to the  
analyzer test port or other static-sensitive device. This can be done as follows:  
1. Connect a short (from your calibration kit) to one end of the cable to short the center  
conductor to the outer conductor.  
2. While wearing a grounded wrist strap, grasp the outer shell of the cable connector.  
3. Connect the other end of the cable to the test port.  
4. Remove the short from the cable.  
Refer to Chapter 6 , Replaceable Parts,” for part numbers and instructions for ordering  
ESD protection devices.  
F igu r e 3-1 E SD P r ote ct ion Se tu p  
3-2  
85054B  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Use, Maintenance, and Care of the Devices  
Visual Inspection  
Visu a l In sp ect ion  
Visual inspection and, if necessary, cleaning should be done every time a connection is  
made. Metal particles from the connector threads may fall into the connector when it is  
disconnected. One connection made with a dirty or damaged connector can damage both  
connectors beyond repair.  
In some cases, magnification is necessary to see damage on a connector; a magnifying  
device with a magnification of 10× is recommended. However, not all defects that are  
visible only under magnification will affect the electrical performance of the connector. Use  
the following guidelines when evaluating the integrity of a connector.  
Look for Obviou s Defects a n d Da m a ge Fir st  
Examine the connectors first for obvious defects and damage: badly worn plating on the  
connector interface, deformed threads, or bent, broken, or misaligned center conductors.  
Connector nuts should move smoothly and be free of burrs, loose metal particles, and  
rough spots.  
Wh a t Ca u ses Con n ect or Wea r ?  
Connector wear is caused by connecting and disconnecting the devices. The more use a  
connector gets, the faster it wears and degrades. The wear is greatly accelerated when  
connectors are not kept clean, or are connected incorrectly.  
Connector wear eventually degrades performance of the device. Calibration devices should  
have a long life if their use is on the order of a few times per week. Replace devices with  
worn connectors.  
The test port connectors on the network analyzer test set may have many connections each  
day, and are therefore more subject to wear. It is recommended that an adapter be used as  
a test port saver to minimize the wear on the test sets test port connectors.  
In sp ect th e Ma tin g P la n e Su r fa ces  
Flat contact between the connectors at all points on their mating plane surfaces is required  
for a good connection. See Figure 2-1 on page 2-3. Look especially for deep scratches or  
dents, and for dirt and metal particles on the connector mating plane surfaces. Also look  
for signs of damage due to excessive or uneven wear or misalignment.  
Light burnishing of the mating plane surfaces is normal, and is evident as light scratches  
or shallow circular marks distributed more or less uniformly over the mating plane  
surface. Other small defects and cosmetic imperfections are also normal. None of these  
affect electrical or mechanical performance.  
If a connector shows deep scratches or dents, particles clinging to the mating plane  
surfaces, or uneven wear, clean and inspect it again. Devices with damaged connectors  
should be discarded. Determine the cause of damage before connecting a new, undamaged  
connector in the same configuration.  
85054B  
3-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
               
Use, Maintenance, and Care of the Devices  
Cleaning Connectors  
In sp ect th e P r ecision Slotless Con n ector s (fem a le)  
Precision slotless female connectors are used to improve accuracy. The slotless contacts are  
not affected by the slight variations in male contact pin diameter. However, it is still  
advisable to inspect them regularly for damage.  
NOTE  
This is particularly important when mating nonprecision to precision devices.  
Clea n in g Con n ect or s  
Clean connectors are essential for ensuring the integrity of RF and microwave coaxial  
connections.  
1. Use Com p r essed Air or Nit r oge n  
WARNING  
Alw a ys u se p r ot ect ive e yew ea r w h en u sin g com p r essed a ir or  
n it r oge n .  
Use compressed air (or nitrogen) to loosen particles on the connector mating plane  
surfaces.  
You can use any source of clean, dry, low-pressure compressed air or nitrogen that has  
an effective oil-vapor filter and liquid condensation trap placed just before the outlet  
hose.  
Ground the hose nozzle to prevent electrostatic discharge, and set the air pressure to  
less than 414 kPa (60 psi) to control the velocity of the air stream. High-velocity streams  
of compressed air can cause electrostatic effects when directed into a connector. These  
electrostatic effects can damage the device. Refer to Electrostatic Discharge” earlier in  
this chapter for additional information.  
2. Clea n th e Con n e ct or Th r ea d s  
WARNING  
Kee p isop r op yl a lcoh ol a w a y fr om h ea t , sp a r k s, a n d fla m e. St or e in a  
tigh tly close d con ta in er. It is ext r em ely fla m m a b le. In ca se of fir e, u se  
a lcoh ol foa m , d r y ch em ica l, or ca r bon d ioxid e; w a t er m a y b e  
in effective.  
Use isop r op yl a lcoh ol w ith a d e qu a te ve n t ila tion a n d a void con ta ct  
w it h eyes, sk in , a n d cloth in g. It ca u ses sk in ir r it a tion , m a y ca u se e ye  
d a m a ge , a n d is h a r m fu l if sw a llow e d or in h a le d . It m a y b e h a r m fu l if  
a bsor b ed th r ou gh t h e sk in . Wa sh th or ou gh ly a ft er h a n d lin g.  
In ca se of sp ill, soa k u p w ith sa n d or ea r th . F lu sh sp ill a r ea w ith  
w a t er.  
Disp ose of isop r op yl a lcoh ol in a ccor d a n ce w ith a ll a p p lica ble  
fed e r a l, st a te , a n d loca l en vir on m en t a l r egu la tion s.  
3-4  
85054B  
Download from Www.Somanuals.com. All Manuals Search And Download.  
             
Use, Maintenance, and Care of the Devices  
Cleaning Connectors  
Use a lint-free swab or cleaning cloth moistened with isopropyl alcohol to remove any  
dirt or stubborn contaminants on a connector that cannot be removed with compressed  
air or nitrogen. Refer to Table 6-2 on page 6-4 for part numbers for isopropyl alcohol and  
cleaning swabs.  
a. Apply a small amount of isopropyl alcohol to a lint-free cleaning swab.  
b. Clean the connector threads.  
c. Let the alcohol evaporate, then blow the threads dry with a gentle stream of clean,  
low-pressure compressed air or nitrogen. Always completely dry a connector before  
you reassemble or use it.  
3. Cle a n t h e Ma t in g P la n e Su r fa ces  
a. Apply a small amount of isopropyl alcohol to a lint-free cleaning swab.  
b. Clean the center and outer conductor mating plane surfaces. Refer to Figure 2-1 on  
page 2-3. When cleaning a female connector, avoid snagging the swab on the center  
conductor contact fingers by using short strokes.  
c. Let the alcohol evaporate, then blow the connector dry with a gentle stream of clean,  
low-pressure compressed air or nitrogen. Always completely dry a connector before  
you reassemble or use it.  
4. In sp ect  
Inspect the connector again to make sure that no particles or residue are present.  
85054B  
3-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Use, Maintenance, and Care of the Devices  
Gaging Connectors  
Ga gin g Con n ect or s  
The gages available from Agilent Technologies are intended for preventive maintenance  
and troubleshooting purposes only. (See Table 6-1 on page 6-2 for part number  
information.) They are effective in detecting excessive center conductor protrusion or  
recession, and conductor damage on DUTs, test accessories, and the calibration kit devices.  
Do not use the gages for precise pin depth measurements.  
Con n ector Ga ge Accu r a cy  
The connector gages are only capable of performing coarse measurements. They do not  
provide the degree of accuracy necessary to precisely measure the pin depth of the kit  
devices. This is partially due to the repeatability uncertainties that are associated with the  
measurement. Only the factory—through special gaging processes and electrical testing—  
can accurately verify the mechanical characteristics of the devices.  
With proper technique, however, the gages are useful in detecting gross pin depth errors on  
device connectors. To achieve maximum accuracy, random errors must be reduced by  
taking the average of at least three measurements having different gage orientations on  
the connector. Even the resultant average can be in error by as much as ± 0.0001 inch due  
to systematic (biasing) errors usually resulting from worn gages and gage masters. The  
information in Table 2-2 on page 2-4 assumes new gages and gage masters. Therefore,  
these systematic errors were not included in the uncertainty analysis. As the gages  
undergo more use, the systematic errors can become more significant in the accuracy of the  
measurement.  
The measurement uncertainties (see Table 2-2 on page 2-4) are primarily a function of the  
assembly materials and design, and the unique interaction each device type has with the  
gage. Therefore, these uncertainties can vary among the different devices. For example,  
note the difference between the uncertainties of the opens and shorts in Table 2-2.  
The observed pin depth limits in Table 2-2 add these uncertainties to the typical factory  
pin depth values to provide practical limits that can be referenced when using the gages.  
See Pin Depth” on page 2-3. Refer to Kit Contents” on page 1-2 for more information on  
the design of the calibration devices in this kit.  
NOTE  
When measuring pin depth, the measured value (resultant average of three  
or more measurements) is not the true value. Always compare the measured  
value with the observed pin depth limits in Table 2-2 on page 2-4 to evaluate  
the condition of device connectors.  
3-6  
85054B  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Use, Maintenance, and Care of the Devices  
Gaging Connectors  
Wh en to Ga ge Con n ector s  
Gage a connector at the following times:  
Prior to using a device for the first time: record the pin depth measurement so that it  
can be compared with future readings. (It will serve as a good troubleshooting tool when  
you suspect damage may have occurred to the device.)  
If either visual inspection or electrical performance suggests that the connector  
interface may be out of typical range (due to wear or damage, for example).  
If a calibration device is used by someone else or on another system or piece of  
equipment.  
Initially after every 100 connections, and after that as often as experience indicates.  
Rea d in g th e Con n ector Ga ge  
The gage dial is divided into increments of 0.0001 inch and major divisions of 0.001 inch  
(see Figure 3-2). For each revolution of the large dial, the smaller dial indicates a change of  
0.01 inch. Use the small dial as the indicator of multiples of 0.01 inch. In most connector  
measuring applications, this value will be zero.  
When making a measurement, the gage dial indicator will travel in one of two directions. If  
the center conductor is recessed from the zero reference plane, the indicator will move  
counterclockwise to indicate the amount of r e cession , which is read as a negative value. If  
the center conductor protrudes, the indicator will move clockwise to indicate the amount of  
p r otr u sion , which is read as a positive value. Refer to Pin Depth” on page 2-3 for  
definitions of protrusion and recession.  
F igu r e 3-2 R ea d in g th e Con n e ct or Ga ge  
85054B  
3-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Use, Maintenance, and Care of the Devices  
Gaging Connectors  
Ga gin g P r oced u r es  
Ga gin g Ma le Typ e-N Con n e ct or s  
NOTE  
Always hold a connector gage by the gage barrel, below the dial indicator.  
This gives the best stability, and improves measurement accuracy.  
1. Select the proper gage for your connector. (Refer to Table 6-2 for gage part numbers).  
2. Inspect and clean the gage, gage master, and device to be gaged. Refer to “Visual  
Inspection” and Cleaning Connectors” earlier in this chapter.  
3. Zero the connector gage (refer to Figure 3-3):  
a. While holding the gage by the barrel, and without turning the gage or the gage  
master, screw the gage master connecting nut onto the male gage, just until you  
meet resistance. Connect the nut finger tight. Do not overtighten.  
b. Use the torque wrench recommended for use with this kit to tighten the connecting  
nut to 135 N-cm (12 in-lb). Refer to Connections” on page 3-16 for more information.  
c. Loosen the dial lock screw on the gage and rotate the gage dial so that the pointer  
corresponds to the correction value noted on the gage master. Do not adjust the gage  
dial to zero, unless the correction value on the gage master is zero.  
d. Tighten the dial lock screw and remove the gage master.  
e. Attach and torque the gage master to the gage once again to verify that the setting is  
repeatable. Remove the gage master.  
4. Gage the device connector (refer to Figure 3-3):  
a. While holding the gage by the barrel, and without turning the gage or the device,  
screw the connecting nut of the device being measured onto the gage, just until you  
meet resistance. Connect the nut finger-tight. Do not overtighten.  
b. Use the torque wrench recommended for use with this kit to tighten the connecting  
nut to 135 N-cm (12 in-lb). Refer to Connections” on page 3-16 for more information.  
c. Gently tap the barrel of the gage with your finger to settle the gage reading.  
d. Read the gage indicator dial. If the needle has moved clockwise, the center conductor  
is protruding by an amount indicated by the black numbers. If the needle has moved  
counterclockwise, the center conductor is recessed by an amount indicated by the red  
numbers.  
For maximum accuracy, measure the connector a minimum of three times and take  
an average of the readings. After each measurement, rotate the gage a quarter-turn  
to reduce measurement variations that result from the gage or the connector face not  
being exactly perpendicular to the center axis.  
e. Compare the average reading with the observed pin depth limits in Table 2-2 on page  
2-4.  
3-8  
85054B  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Use, Maintenance, and Care of the Devices  
Gaging Connectors  
F igu r e 3-3 Ga gin g Ma le Typ e-N Con n ector s  
85054B  
3-9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Use, Maintenance, and Care of the Devices  
Gaging Connectors  
Ga gin g Fem a le Typ e-N Con n e ct or s  
NOTE  
Always hold a connector gage by the gage barrel, below the dial indicator.  
This gives the best stability, and improves measurement accuracy.  
1. Select the proper gage for your connector. (Refer to Table 6-2 for gage part numbers).  
2. Inspect and clean the gage, gage master, and device to be gaged. Refer to “Visual  
Inspection” and Cleaning Connectors” earlier in this chapter.  
3. Zero the connector gage (refer to Figure 3-4):  
a. While holding the gage by the barrel, and without turning the gage or the gage  
master, screw the gage connecting nut onto the female gage master, just until you  
meet resistance. Connect the nut finger-tight. Do not overtighten.  
b. Use the torque wrench recommended for use with this kit to tighten the connecting  
nut to 135 N-cm (12 in-lb). Refer to Connections” on page 3-16 for more information.  
c. Loosen the dial lock screw on the gage and rotate the gage dial so that the pointer  
corresponds to the correction value noted on the gage master. Do not adjust the gage  
dial to zero, unless the correction value on the gage master is zero.  
d. Tighten the dial lock screw and remove the gage master.  
e. Attach and torque the gage master to the gage once again to verify that the setting is  
repeatable. Remove the gage master.  
4. Gage the device connector (refer to Figure 3-3 on page 3-9):  
a. While holding the gage by the barrel, and without turning the gage or the device,  
screw the gage connecting nut onto the device being measured, just until you meet  
resistance. Connect the nut finger-tight. Do not overtighten.  
b. Use the torque wrench recommended for use with this kit to tighten the connecting  
nut to 135 N-cm (12 in-lb). Refer to Connections” on page 3-16 for more information.  
c. Gently tap the barrel of the gage with your finger to settle the gage reading.  
d. Read the gage indicator dial. If the needle has moved clockwise, the center conductor  
is protruding by an amount indicated by the black numbers. If the needle has moved  
counterclockwise, the center conductor is recessed by an amount indicated by the red  
numbers.  
For maximum accuracy, measure the connector a minimum of three times and take  
an average of the readings. After each measurement, rotate the gage a quarter-turn  
to reduce measurement variations that result from the gage or the connector face not  
being exactly perpendicular to the center axis.  
e. Compare the average reading with the observed pin depth limits in Table 2-2 on page  
2-4.  
3-10  
85054B  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Use, Maintenance, and Care of the Devices  
Gaging Connectors  
F igu r e 3-4 Ga gin g Fe m a le Typ e-N Con n ector s  
85054B  
3-11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Use, Maintenance, and Care of the Devices  
Gaging Connectors  
Ga gin g t h e Slid in g Loa d  
Gage the sliding load before each use. If the sliding load pin depth is out of the observed  
pin depth limits listed in Table 2-2 on page 2-4, refer to Adjusting the Sliding Load Pin  
NOTE  
Always hold a connector gage by the gage barrel, below the dial indicator.  
This gives the best stability, and improves measurement accuracy. (Cradling  
the gage in your hand or holding it by the dial applies stress to the gage  
plunger mechanism through the dial indicator housing.)  
NOTE  
The sliding load uses a plastic centering bead to support its center conductor  
when pin depth is adjusted and gaged and when the load is stored. Remove  
this support bead from the sliding load before you connect the load for an  
electrical calibration. Reinsert this support bead when youve finished using  
the sliding load.  
1. Select the proper gage for your connector. Refer to Replaceable Parts for the 85054B  
Calibration Kit” on page 6-2 for gage part numbers.  
2. Inspect and clean the gage, gage master, and device to be gaged. Refer to “Visual  
3. Zero the connector gage as described in either Step 3 on page 3-8 (for a male gage) or  
Step 3 on page 3-10 (for a female gage)  
4. Remove the center conductor protective cap from the sliding load.  
5. Loosen the center conductor pull-back nut completely, and press the center-conductor  
cap to extend the center conductor beyond the end of the connector. With the sliding  
ring pulled back approximately 0.5 inch, install a centering bead (if not already  
installed) in the connector end of the sliding load.  
6. Continue to press the center conductor cap and mate the center conductor of the sliding  
load with the gages center conductor.  
CAUTION  
The sliding load center conductor can be damaged if the sliding load is not  
held in line when mating the load to a connector. Always line up the sliding  
load when connecting or removing it from a connector.  
7. Mate the outer conductor of the sliding load with the outer conductor of the gage.  
Torque the connection with a 3/4 inch torque wrench to approximately 135 N-cm  
(12 in-lb). Retighten the center conductor pull-back nut. It will click” when it is tight.  
8. Gently tap the barrel of the gage with your finger to settle the gage reading.  
9. Read the gage indicator dial. If the needle had moved clockwise, the center conductor is  
protruding and the value is determined by the black numbers. If the needle had moved  
counterclockwise, the center conductor is recessed by an amount determined by the red  
numbers.  
10.For maximum accuracy, measure the connector a minimum of three times and take an  
average of the readings.  
3-12  
85054B  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Use, Maintenance, and Care of the Devices  
Gaging Connectors  
NOTE  
When performing pin depth measurements, use different orientations of the  
gage within the connector. Averaging a minimum of three readings, each  
taken after a quarter-turn rotation of the gage, reduces measurement  
variations that result from the gage or the connector face not being exactly  
perpendicular to the center axis.  
11.Compare the average reading with the observed pin depth limits in Table 2-2 on page  
2-4. If the pin depth is outside the limits, follow the procedure Adjusting the Sliding  
12.Loosen the connection between the gage and the sliding load, and remove the sliding  
load from the gage.  
13.Leave the centering bead installed on the sliding load if you are going to adjust the pin  
depth. Carefully remove the centering bead from the sliding load if youre going to use it  
for an electrical calibration. If the centering bead does not come out of the sliding load  
easily, loosen the center conductor pull-back nut, and press the center conductor cap to  
extend the center conductor. This should expose the centering bead so that it may be  
removed. Retract the center conductor and retighten the pull-back nut.  
If the centering bead still will not come out, hold the sliding load with the connector end  
pointed down. Move the sliding element up, then quickly down. The trapped air behind  
the centering bead helps eject it.  
CAUTION  
Damage can occur to the sliding load during the removal of a centering bead  
that has slipped too far into the sliding load. If youre going to perform an  
electrical calibration, prevent damage by removing the centering bead  
immediately after gaging the sliding load pin depth. The sliding load will not  
perform to its specifications if the centering bead is not removed from the  
sliding load before an electrical calibration.  
F igu r e 3-5 Ga gin g th e Slid in g Loa d P in Dep t h  
Ad ju stin g th e Slid in g Loa d P in Dep th  
The sliding loads in this kit have a setback mechanism that allows the pin depth to be set  
to any desired value. The pin depth of the sliding load is preset at the factory. The pin  
depth should not ha ve to be reset ea ch time the sliding load is used, but it should be checked  
before each use.  
85054B  
3-13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Use, Maintenance, and Care of the Devices  
Gaging Connectors  
If the pin depth is outside the observed limits listed in Table 2-2 on page 2-4, use the  
following procedure to reset it. Always measure the sliding load pin depth before attaching  
it to any connector.  
This procedure assumes that you were directed here from Gaging the Sliding Load” on  
page 3-12. If not, perform the steps in that procedure before performing this procedure.  
1. The gage should be attached to the sliding load. The sliding load should have its  
centering bead installed. Refer to Gaging the Sliding Load” on page 3-12 if necessary.  
2. With a 0.050 inch hex key, loosen the two la rgest hex screws by turning them 1/4 turn  
clockwise. Refer to Figure 3-6.  
CAUTION  
Do not loosen any hex screws other than the two largest hex screws pointed  
out in Figure 3-6.  
3. Gently turn the center conductor pin depth adjustment knob on the sliding load until  
the gage pointer reads 3.81 micrometers (1.5 x 10–4 inches). Refer to Figure 3-6.  
4. Tighten the two hex screws just until they are finger tight (do not overtighten).  
5. Wait approximately five minutes to allow the temperature to stabilize. Do not touch  
either the gage or the sliding load during this time.  
6. Note the gage reading. If it is no longer within the allowable range, perform steps 2–5  
again.  
7. Loosen the connecting nut and remove the gage from the sliding load. If youre going to  
store the sliding load, leave the centering bead installed. If youre gong to use the  
sliding load for an electrical calibration, remove the centering bead.  
NOTE  
When performing pin depth measurements, use different orientations of the  
gage within the connector. Averaging a minimum of three readings, each  
taken after a quarter-turn rotation of the gage, reduces measurement  
variations that result from the gage or the connector face not being exactly  
perpendicular to the center axis.  
The sliding load pin depth is now is specification and the load is ready to use. Once the  
sliding load pin depth is set it rarely needs to be adjusted. However, the pin depth should  
be rechecked before each use. Replace the protective plastic caps on the sliding load and  
gage connectors when these devices are not in use.  
3-14  
85054B  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Use, Maintenance, and Care of the Devices  
Gaging Connectors  
F igu r e 3-6 Ad ju stin g th e Slid in g Loa d P in Dep t h  
85054B  
3-15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Use, Maintenance, and Care of the Devices  
Connections  
Con n ection s  
Good connections require a skilled operator. The most common cause of measurement error  
is bad connections. The following procedures illustrate how to make good connections.  
How to Ma k e a Con n ection  
P r e lim in a r y Con n ection  
1. Ground yourself and all devices. Wear a grounded wrist strap and work on a grounded,  
conductive table mat. Refer to Electrostatic Discharge” on page 3-2 for ESD  
precautions.  
2. Visually inspect the connectors. Refer to “Visual Inspection” on page 3-3.  
3. If necessary, clean the connectors. Refer to Cleaning Connectors” on page 3-4.  
4. Use a connector gage to verify that all center conductors are within the observed pin  
5. Carefully align the connectors. The male connector center pin must slip concentrically  
into the contact finger of the female connector.  
6. Push the connectors straight together.  
CAUTION  
Do not turn the device body. Only turn the connector nut. Damage to the  
center conductor can occur if the device body is twisted.  
Do not twist or screw the connectors together. As the center conductors mate, there is  
usually a slight resistance.  
7. The preliminary connection is tight enough when the mating plane surfaces make  
uniform, light contact. Do not overtighten this connection.  
A connection in which the outer conductors make gentle contact at all points on both  
mating surfaces is sufficient. Very light finger pressure is enough to accomplish this.  
8. Make sure the connectors are properly supported. Relieve any side pressure on the  
connection from long or heavy devices or cables.  
F in a l Con n ection Usin g a Tor q u e Wr e n ch  
Use a torque wrench to make a final connection. Table 3-1 provides information about the  
torque wrench recommended for use with this calibration kit. A torque wrench is not  
included in the calibration kit. Refer to Chapter 6 Replaceable Parts” for part number and  
ordering information.  
Ta b le 3-1 Tor q u e Wr e n ch In for m a t ion  
Connector Type  
Torque Setting  
Torque Tolerance  
Type-N  
135 N-cm (12 in-lb)  
±13.5 N-cm (±1.2 in-lb)  
3-16  
85054B  
Download from Www.Somanuals.com. All Manuals Search And Download.  
                 
Use, Maintenance, and Care of the Devices  
Connections  
Using a torque wrench guarantees that the connection is not too tight, preventing possible  
connector damage. It also guarantees that all connections are equally tight each time.  
Prevent the rotation of anything other than the connector nut that you are tightening. It  
may be possible to do this by hand if one of the connectors is fixed (as on a test port). In all  
situations, however, it is recommended that you use an open-end wrench to keep the body  
of the device from turning. Refer to Chapter 6 for part number and ordering information.  
1. Position both wrenches within 90 degrees of each other before applying force. See  
Figure 3-7. Wrenches opposing each other (greater than 90 degrees apart) will cause a  
lifting action which can misalign and stress the connections of the devices involved.  
This is especially true when several devices are connected together.  
F igu r e 3-7 Wr en ch P osition s  
2. Hold the torque wrench lightly, at the end of the handle only (beyond the groove). See  
F igu r e 3-8 Usin g th e Tor q u e Wr e n ch  
85054B  
3-17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Use, Maintenance, and Care of the Devices  
Connections  
3. Apply downward force perpendicular to the wrench handle. See Figure 3-8. This applies  
torque to the connection through the wrench.  
Do not hold the wrench so tightly that you push the handle straight down along its  
length rather than pivoting it, otherwise you apply an unknown amount of torque.  
4. Tighten the connection just to the torque wrench break point. The wrench handle gives  
way at its internal pivot point. See Figure 3-8. Do not tighten the connection further.  
CAUTION  
You dont have to fully break the handle of the torque wrench to reach the  
specified torque; doing so can cause the handle to kick back and loosen the  
connection. Any give at all in the handle is sufficient torque.  
Do not pivot the wrench handle on your thumb or other fingers, otherwise you apply an  
unknown amount of torque to the connection when the wrench reaches its break point.  
Do not twist the head of the wrench relative to the outer conductor mating plane. If you  
do, you apply more than the recommended torque.  
Con n ectin g th e Slid in g Loa d  
Use this procedure to connect the sliding load to a test port or a type-N cable connector.  
NOTE  
The sliding load uses a plastic centering bead to support its center conductor  
when pin depth is adjusted and gaged and when the load is stored. Remove  
this support bead from the sliding load before you connect the load for an  
electrical calibration. Reinsert this support bead when youve finished using  
the sliding load.  
CAUTION  
CAUTION  
Circuitry inside the test set at the test ports may be destroyed if precautions  
are not taken to avoid electrostatic discharge (ESD). During this procedure,  
the center conductor of the sliding load is connected to the exposed center  
conductor of the test port. Ground yourself to prevent electrostatic discharge.  
The sliding load center conductor can be damaged if the sliding load is not  
held in line when mating the load to a connector. Always line up the sliding  
load when connecting or removing it from a connector.  
1. Refer to Figure 3-9. Loosen the center conductor pull-back nut completely. Press the  
center conductor cap to extend the center conductor of the sliding load beyond the end of  
the connector.  
2. Continue to press the center conductor cap and mate the center conductor of the sliding  
load with the cable/test port connectors center conductor.  
3-18  
85054B  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Use, Maintenance, and Care of the Devices  
Connections  
3. Release pressure on the center conductor and mate the outer conductor of the sliding  
load with the outer conductor of the cable/test port connector. Torque the connection  
with a 3/4 inch torque wrench to approximately 135 N-cm (12 in-lb). Refer to Final  
4. Retighten the center conductor pull-back nut. It will click” when it is tight.  
F igu r e 3-9 Con n ect in g t h e Slid in g Loa d  
How to Sep a r a te a Con n ection  
To avoid lateral (bending) force on the connector mating plane surfaces, always support the  
devices and connections.  
CAUTION  
Turn the connector nut, not the device body. Major damage to the center  
conductor can occur if the device body is twisted.  
1. Use an open-end wrench to prevent the device body from turning.  
2. Use another open-end wrench to loosen the connector nut.  
3. Complete the separation by hand, turning only the connector nut.  
4. Pull the connectors straight apart without twisting, rocking, or bending either of the  
connectors.  
85054B  
3-19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Use, Maintenance, and Care of the Devices  
Using the Sliding Load  
Usin g t h e Slid in g Loa d  
When performing a sliding load calibration, it is recommended that the sliding ring be set  
at the marked positions (rings) along the sliding load body. Using the set marks ensures  
that a broad distribution of phase angles is selected, thereby optimizing the calibration.  
The set marks function as detents so that the internal center of the sliding ring can mate  
with them. Because of this, the set mark being used cannot be seen but is felt as the sliding  
ring is moved from mark to mark during a calibration. Moving the sliding ring with only  
the index fingers of both hands will increase your ability to detect the sliding ring detent at  
each position.  
To perform a sliding load calibration, refer to your network analyzers users documentation  
for instructions.  
3-20  
85054B  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Use, Maintenance, and Care of the Devices  
Using the Sliding Load  
F igu r e 3-10 Usin g th e Slid in g Loa d  
85054B  
3-21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Use, Maintenance, and Care of the Devices  
Handling and Storage  
Ha n d lin g a n d St or a ge  
Install the protective end caps and store the calibration devices in the foam-lined  
storage case when not in use.  
Never store connectors loose in a box, desk, or bench drawer. This is the most common  
cause of connector damage during storage.  
Keep connectors clean.  
Do not touch mating plane surfaces. Natural skin oils and microscopic particles of dirt  
are easily transferred to a connector interface and are very difficult to remove.  
Do not set connectors contact-end down on a hard surface. The plating and the mating  
plane surfaces can be damaged if the interface comes in contact with any hard surface.  
3-22  
85054B  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
4 P er for m a n ce Ver ifica t ion  
4-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Performance Verification  
Introduction  
In t r od u ction  
The performance of your calibration kit can only be verified by returning the kit to Agilent  
Technologies for recertification. The equipment required to verify the specifications of the  
devices in the kit has been specially manufactured and is not commercially available.  
How Agilen t Ver ifies th e Devices in Th is Kit  
Agilent verifies the specifications of these devices as follows:  
1. The residual microwave error terms of the test system are verified with precision  
airlines and shorts that are directly traced to NIST (National Institute of Standards  
and Technology). The airline and short characteristics are developed from mechanical  
measurements. The mechanical measurements and material properties are carefully  
modeled to give very accurate electrical representation. The mechanical measurements  
are then traced to NIST through various plug and ring gages and other mechanical  
measurements.  
2. Each calibration device is electrically tested on this system. For the initial (before sale)  
testing of the calibration devices, Agilent includes the test measurement uncertainty as  
a guardband to guarantee each device meets the published specification. For  
recertifications (after sale), no guardband is used and the measured data is compared  
directly with the specification to determine the pass or fail status. The measurement  
uncertainty for each device is, however, recorded in the calibration report that  
accompanies recertified kits.  
These two steps establish a traceable link to NIST for Agilent to the extent allowed by the  
institutes calibration facility. The specifications data provided for the devices in this kit is  
traceable to NIST through Agilent Technologies.  
4-2  
85054B  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Performance Verification  
Recertification  
R ecer t ifica t ion  
The following will be provided with a recertified kit:  
a new calibration sticker affixed to the case  
a certificate of calibration  
a calibration report for each device in the kit listing measured values, specifications,  
and uncertainties  
NOTE  
A list of NIST traceable numbers may be purchased upon request to be  
included in the calibration report.  
Agilent Technologies offers a Sta ndard calibration for the recertification of this kit. For  
more information, contact Agilent Technologies. For contact information, see page 5-3.  
How Often to R ecer tify  
The suggested initial interval for recertification is 12 months or sooner. The actual need for  
recertification depends on the use of the kit. After reviewing the results of the initial  
recertification, you may establish a different recertification interval that reflects the usage  
and wear of the kit.  
NOTE  
The recertification interval should begin on the date the kit is first used after  
the recertification date.  
Wh er e to Sen d a Kit for R ecer tifica t ion  
Contact Agilent Technologies for information on where to send your kit for recertification.  
For contact information, refer to page 5-3.  
When you return the kit, complete and attach a service tag. Refer to Returning a Kit or  
85054B  
4-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
           
Performance Verification  
Recertification  
4-4  
85054B  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5 Tr ou blesh ootin g  
5-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Troubleshooting  
Troubleshooting Process  
Tr ou b lesh oot in g P r ocess  
If you suspect a bad calibration, or if your network analyzer does not pass performance  
verification, follow the steps in Figure 5-1.  
F igu r e 5-1 Tr ou b lesh oot in g F low ch a r t  
5-2  
85054B  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Troubleshooting  
Returning a Kit or Device to Agilent  
R et u r n in g a Kit or Device t o Agilen t  
If your kit or device requires service, contact Agilent Technologies for information on  
where to send it. See Table 5-1 for contact information. Include a service tag (located near  
the end of this manual) on which you provide the following information:  
your company name and address  
a technical contact person within your company, and the person's complete telephone  
number  
the model number and serial number of the kit  
the part number and serial number of each device  
the type of service required  
a deta iled description of the problem and how the device was being used when the  
problem occurred (such as calibration or measurement)  
Wh er e to Look for Mor e In for m a tion  
This manual contains limited information about network analyzer system operation. For  
complete information, refer to the instrument documentation. If you need additional  
information, contact Agilent Technologies.  
Con t a ct in g Agilen t  
Ta b le 5-1 Con ta ctin g Agilen t  
On lin e a ssist a n ce: www.agilent.com/find/assist  
Un it e d Sta t e s  
La t in Am e r ica  
Ca n a d a  
Eu r op e  
(tel) 1 800 452 4844  
(tel) (305) 269 7500  
(fax) (305) 269 7599  
(tel) 1 877 894 4414  
(fa x) (905) 282-6495  
(tel) (+31) 20 547 2323  
(fax) (+31) 20 547 2390  
New Ze a la n d  
J a p a n  
Au st r a lia  
Sin ga p or e  
(tel) 0 800 738 378  
(fax) (+64) 4 495 8950  
(tel) (+81) 426 56 7832  
(fax) (+81) 426 56 7840  
(tel) 1 800 629 485  
(fa x) (+61) 3 9210 5947 (fax) (65) 836 0252  
(tel) 1 800 375 8100  
Ma la ysia  
P h ilip p in es  
Th a ila n d  
Hon g Kon g  
(tel) 1 800 828 848  
(fax) 1 800 801 664  
(tel) (632) 8426802  
(tel) (PLDT subscriber  
only):  
1 800 16510170  
(fax) (632) 8426809  
(fax) (PLDT subscriber  
only):  
(tel) outside Bangkok:  
(088) 226 008  
(tel) within Ba ngkok:  
(662) 661 3999  
(tel) 800 930 871  
(fax) (852) 2506 9233  
(fa x) (66) 1 661 3714  
1 800 16510288  
Ta iw a n  
(tel) 0800-047-866  
(fax) (886) 2 25456723  
P eop le s R ep u blic of  
Ch in a  
(tel) (preferred):  
800-810-0189  
In d ia  
(tel) 1-600-11-2929  
(fa x) 000-800-650-1101  
(tel) (alternate):  
10800-650-0021  
(fax) 10800-650-0121  
85054B  
5-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
               
Troubleshooting  
Returning a Kit or Device to Agilent  
5-4  
85054B  
Download from Www.Somanuals.com. All Manuals Search And Download.  
6 R ep la cea ble Pa r ts  
6-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Replaceable Parts  
Introduction  
In t r od u ction  
Table 6-1 lists the replacement part numbers for items included in the 85054B calibration  
kit and Figure 6-1 illustrates each of these items.  
Table 6-2 lists the replacement part numbers for items recommended or required for  
successful operation but not included in the calibration kit.  
To order a listed part, note the description, the part number, and the quantity desired.  
Ta b le 6-1 R e p la cea b le P a r ts for t h e 85054B Ca lib r a tion Kit  
Item No.  
Description  
Qty Per  
Kit  
Agilent Part Number  
Ca libr a t ion Devices (50Typ e-N)  
1
2
3
4
5
6
7
8
Type-N (m) sliding load  
Type-N (f) sliding load  
Type-N (m) lowband load  
Type-N (f) lowband load  
Type-N (m) offset short  
Type-N (f) offset short  
Type-N (m) offset open  
Type-N (f) offset open  
1
1
1
1
1
1
1
1
85054-60035  
85054-60036  
00909-60011  
00909-60012  
85054-60025  
85054-60026  
85054-60027  
85054-60028  
Ad a p t er s  
9
Type-N (m) to Type-N (m)  
Type-N (f) to Type-N (f)  
Type-N (f) to 7 mm  
1
1
2
2
85054-60038  
85054-60037  
85054-60031  
85054-60032  
10  
11  
12  
Type-N (m) to 7 mm  
Wr en ch es  
3/4 in., 135 N-cm (12 in-lb) Torque  
Spanner  
13  
14  
1
1
8710-1766  
08513-20014  
Ca libr a t ion Kit St or a ge Ca se  
15  
16  
17  
18  
19  
Storage Case Assembly (includes items listed below)  
1
1
1
1
1
85054-60039  
5180-7900  
Box  
ID label  
85054-80012  
85054-80023  
5181-5543  
Foam pad (bottom)  
Foam pad (lid)  
6-2  
85054B  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Replaceable Parts  
Introduction  
Ta b le 6-1 R ep la cea b le P a r t s for th e 85054B Ca libr a t ion Kit  
Item No.  
Description  
Qty Per  
Kit  
Agilent Part Number  
20  
Disk holder  
1
5180-8491  
Ga ges  
21  
22  
23  
24  
25  
26  
27  
Screw-on N Gage  
1
1
1
1
1
1
2
85054-80011  
85054-60049  
85054-60050  
85054-60052  
85054-60051  
85054-60053  
85054-80028  
Type-N gage set (includes items listed below)  
Type-N gage (f)  
Type-N gage master (f)  
Type-N gage (m)  
Type-N gage master (m)  
Centering beads  
Miscella n e ou s Ite m s  
28  
29  
30  
31  
Users and service guide  
1
1
1
1
85054-90049  
85054-10007  
85054-10005  
85010-10033  
Calibration definitions disk (PNA)  
Calibration definitions disk (8510, 872x series)  
a
Specifications & Performance Verification Disk  
Protective End Cap (f)  
32  
33  
34  
as  
1401-0225  
1401-0208  
08510-90360  
required  
Protective End Cap (m & 7 mm)  
as  
required  
Connector Care–Quick Reference Card  
1
a. See the 8510C On-Site Service Ma nual for instructions on using this disk.  
85054B  
6-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Replaceable Parts  
Introduction  
Ta b le 6-2 R ep la ce a ble P a r tsIt em s Not In clu d ed in t h e Ca lib r a tion Kit  
Item No.  
Description  
Qty  
Agilent Part Number  
Wr e n ch es  
1
2
1/2 in and 9/16 in open-end wrench  
3/4 in open-end wrench  
1
1
8710-1770  
8720-0011  
Ad a p t er s  
3
4
50 ohm (m) Type-N to 7 mm (extendable/retractable sleeve)  
50 ohm (f) Type-N to 7 mm (extendable/retractable sleeve)  
E SD P r ot e ct ive Devices  
1
1
85054-60009  
85054-60001  
5
6
7
8
Grounding wrist strap  
1
1
1
1
9300-1367  
9300-0980  
9300-0797  
9300-1126  
5 ft grounding cord for wrist strap  
2 ft by 4 ft conductive table mat with 15 ft grounding wire  
ESD heel strap  
Con n e ct or Clea n in g Su p p lies  
Isopropyl alcohol  
9
30 ml 8500-5344  
100 9301-1243  
10  
Foam tipped cleaning swabs  
6-4  
85054B  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Replaceable Parts  
Introduction  
F igu r e 6-1 R ep la cea b le P a r t s for th e 85054B Ca libr a tion Kit  
85054B  
6-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Replaceable Parts  
Introduction  
F igu r e 6-2 Mor e R ep la ce a ble P a r ts for t h e 85054B Ca lib r a tion Kit  
6-6  
85054B  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
A Sta n d a r d Defin ition s  
85054B  
A-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Standard Definitions  
Version Changes  
Ver sion Ch a n ges  
Class assignments and standard definitions may change as more accurate model and  
calibration methods are developed. The disk shipped with the kit will contain the most  
recent version.  
A-2  
85054B  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Standard Definitions  
Standard Class Assignments  
St a n d a r d Cla ss Assign m en ts  
Class assignment organizes calibration standards into a format compatible with the error  
models used in the measurement calibration. A class or group of classes corresponds to the  
systematic errors to be removed from the measured network analyzer response. Table A-1  
through A-3 list the classes of the devices in the kit for various network analyzers. This  
information resides on the calibration definitions disk included in the kit.  
Ta b le A-1 St a n d a r d Cla ss Assign m en t s for th e 8510 Net w or k An a lyzer  
Disk File Name: CK_NTYPB2  
Calibration Kit Label: TYPE N B.2  
File Number: *FILE 1  
Class  
A
B
C
12  
12  
D
E
F
G
Standard Class Label  
S
S
S
S
S
S
A
B
C
A
B
C
2
1
9
2
1
9
4
3
Open  
Short  
Loads  
Open  
Short  
Loads  
11  
11  
11  
22  
22  
22  
10  
4
3
10  
Forward transmission  
Reverse transmission  
Forward match  
11  
11  
11  
11  
9
Thru  
Thru  
Thru  
Reverse match  
Thru  
a
Isoln Std  
Forward isolation  
Reverse isolation  
Frequency response  
TRL thru  
9
Isoln Std  
Response  
Undefined  
Undefined  
Undefined  
Adapter  
1
3
2
4
11  
14  
1
TRL reflect  
TRL line  
15  
13  
Adapter  
14  
TRL Option  
Cal Z :  
_____ System Z  
__X__ Thru  
__X__ Line Z  
0
0
0
Set ref:  
_____ Reflect  
Lowband frequency: 2.0 GHz  
a. The forward isolation standard is also used for the isolation part of the response and isolation  
calibration.  
85054B  
A-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Standard Definitions  
Standard Class Assignments  
Ta b le A-2 Sta n d a r d Cla ss Assign m en ts for t h e 872x Ser ies Netw or k An a lyzer  
Calibration Kit Label: [N 50 ]  
Cla ss  
A
B
C
D
E
F
G
St a n d a r d Cla ss La b el  
2
1
3
2
1
3
8
7
5
8
7
5
Open  
Short  
Loads  
Open  
Short  
Loads  
S
S
S
S
S
S
A
B
C
A
B
C
11  
11  
11  
22  
22  
22  
6
6
Forward transmission  
Reverse transmission  
Forward match  
Reverse match  
Response  
4
4
4
4
1
1
4
2
3
Thru  
Thru  
Thru  
Thru  
7
7
2
2
8
8
4
4
Response  
Response  
Thru  
Response & isolation  
TRL thru  
TRL reflect  
8
5
Open  
TRL line or match  
6
Loads  
TRL Op t ion  
__X__ Line Z  
Cal Z :  
_____ System Z  
0
0
0
Set ref:  
__X__ Thru  
_____ Reflect  
A-4  
85054B  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Standard Definitions  
Standard Class Assignments  
Ta b le A-3 St a n d a r d Cla ss Assign m en t s for th e P NA Ser ies Netw or k An a lyzer  
Ca lib r a t ion Kit La b el:  
Type-N Model 85054B  
Aa  
Cla ss  
2
S
S
S
S
S
S
S
S
A
B
C
T
A
B
C
T
11  
11  
11  
21  
22  
22  
22  
12  
1
3, 5, 6  
4
2
1
3, 5, 6  
4
a. For additional ports, make sure  
values match the correct sex of  
the port.  
IMPORTANT The following calibrations are only supported by certain PNA analyzers. See  
your PNA series network analyzer embedded help system.  
Notes:  
1. If you a r e p er for m in g a TR L ca lib r a tion :  
S
S
21T and S12T must be defined as thru standards.  
11A and S22A must be defined as reflection standards.  
S11B, S11C, S22B, and S22C must be defined as line standards.  
2. If you a r e p er for m in g a TR M ca lib r a tion :  
S21T and S12T must be defined as thru standards.  
11A and S22A must be defined as reflection standards.  
S11B, S11C, S22B, and S22C must be defined as ma tch standards.  
S
3. If you a r e p er for m in g a n LR M ca libr a tion :  
S21T and S12T must be defined as line standards.  
S11A and S22A must be defined as reflection standards.  
S
11B, S11C, S22B, and S22C must be defined as ma tch standards.  
85054B  
A-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Standard Definitions  
Standard Class Assignments  
4. S11B a n d S11C m u st b e d efin e d a s t h e sa m e sta n d a r d .  
5. S22B a n d S22C m u st b e d efin e d a s t h e sa m e sta n d a r d .  
For additional information on performing TRL, TRM, and LRM calibrations, refer to your  
PNA series network analyzer embedded help system.  
A-6  
85054B  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Standard Definitions  
Standard Class Assignments  
Bla n k For m s  
The standard class assignments may be changed to meet your specific requirements. Table  
A-4 through A-6 are provided to record the modified standard class assignments.  
Ta b le A-4  
Bla n k For m for th e 8510 Netw or k An a lyzer  
Ca libr a t ion Kit La bel: _________________  
Disk File Name: _______________________  
File Number:  
Cla ss  
A
B
C
D
E
F
G
St a n d a r d Cla ss La bel  
S
S
S
S
S
S
A
B
C
A
B
C
11  
11  
11  
22  
22  
22  
Forward transmission  
Reverse transmission  
Forward match  
Reverse match  
a
Forward isolation  
Reverse isolation  
Frequency response  
TRL thru  
TRL reflect  
TRL line  
Adapter  
TR L Op t ion  
_____ Line Z  
Cal Z :  
_____ System Z  
0
0
0
Set ref:  
_____ Thru  
_____ Reflect  
b
Lowband frequency :  
___________  
a. The forward isolation standard is also used for the isolation part of the response and isolation calibration.  
b. Broadband loads are used for frequencies up to 2 GHz.  
85054B  
A-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Standard Definitions  
Standard Class Assignments  
Ta b le A-5  
Bla n k For m for t h e 872x Se r ie s of Net w or k An a lyze r s  
Ca lib r a t ion Kit La be l: _________________  
St a n d a r d Cla ss La be l  
Cla ss  
A
B
C
D
E
F
G
S
S
S
S
S
S
A
B
C
A
B
C
11  
11  
11  
22  
22  
22  
Forward transmission  
Reverse transmission  
Forward match  
Reverse match  
Response  
Response & isolation  
TRL thru  
TRL reflect  
TRL line or match  
TR L Op t ion  
_____ Line Z  
Cal Z :  
_____ System Z  
_____ Thru  
0
0
0
Set ref:  
_____ Reflect  
A-8  
85054B  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Standard Definitions  
Standard Class Assignments  
Ta b le A-6 Bla n k For m for th e P NA Se r ie s Net w or k An a lyze r s  
Ca lib r a t ion Kit La b el:  
________________________  
Aa  
Cla ss  
S
S
S
S
S
S
S
S
A
B
C
T
A
B
C
T
11  
11  
11  
21  
22  
22  
22  
12  
a. For additional ports, make sure  
values match the correct sex of  
the port.  
IMPORTANT The following calibrations are only supported by certain PNA analyzers. See  
your PNA series network analyzer embedded help system.  
Notes:  
1. If you a r e p er for m in g a TR L ca lib r a tion :  
S
S
21T and S12T must be defined as thru standards.  
11A and S22A must be defined as reflection standards.  
S11B, S11C, S22B, and S22C must be defined as line standards.  
2. If you a r e p er for m in g a TR M ca lib r a tion :  
S21T and S12T must be defined as thru standards.  
11A and S22A must be defined as reflection standards.  
S11B, S11C, S22B, and S22C must be defined as ma tch standards.  
S
3. If you a r e p er for m in g a n LR M ca libr a tion :  
S21T and S12T must be defined as line standards.  
S11A and S22A must be defined as reflection standards.  
S
11B, S11C, S22B, and S22C must be defined as ma tch standards.  
85054B  
A-9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Standard Definitions  
Standard Class Assignments  
4. S11B a n d S11C m u st b e d efin e d a s t h e sa m e sta n d a r d .  
5. S22B a n d S22C m u st b e d efin e d a s t h e sa m e sta n d a r d .  
For additional information on performing TRL, TRM, and LRM calibrations, refer to your  
PNA series network analyzer embedded help system.  
A-10  
85054B  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Standard Definitions  
Nominal Standard Definitions  
Nom in a l St a n d a r d Defin it ion s  
Standard definitions provide the constants needed to mathematically model the electrical  
characteristics (delay, attenuation, and impedance) of each calibration standard. The  
nominal values of these constants are theoretically derived from the physical dimensions  
and material of each calibration standard, or from actual measured response. These values  
are used to determine the measurement uncertainties of the network analyzer. The  
standard definitions in Table A-7 through A-9 list typical calibration kit parameters used  
to specify the mathematical model of each device. This information must be loaded into the  
network analyzer to perform valid calibrations. Refer to your network analyzer users  
guide for instructions on loading calibration definitions.  
NOTE  
The values in the standard definitions table are valid only over the specified  
operating temperature range.  
Sett in g th e System Im p ed a n ce  
This kit contains only 50 ohm devices. Ensure the system impedance (Z0) is set to 50 ohms.  
Refer to your network analyzers users guide for instructions on setting system impedance.  
85054B  
A-11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Standard Definitions  
Nominal Standard Definitions  
Ta b le A-7  
St a n d a r d Defin it ion s for th e 8510 Ne tw or k An a lyzer  
System Z0a = 50.0 Ω  
Disk File Name: CK_NTYPB2  
Calibration Kit Label: TYPE N B.2  
File Number: * FILE 1  
Fr equ en cy  
Sta n d a r d b  
in GH zd  
Offset  
Shorte  
Short (m)f  
1
2
0.1315 606.21 −68.405 2.0206  
104.13 1943.4 144.62  
2.2258  
27.990  
22.905  
50 1.3651  
50 0.93  
0
0
999  
999  
Coax  
Coax  
Open  
Open (m)  
3
4
0.7563  
89.939  
459.88 52.429 1.5846  
2536.8 264.99 13.4  
63.078  
57.993  
50 1.1273  
50 0.93  
0
0
999  
999  
Coax  
Coax  
Short  
Short (f)  
Open  
Open (f)  
5
6
7
8
9
Load  
Load  
Fixed  
0
0
0
50  
50  
50  
0
0
0
0
1.999  
0
999  
999  
999  
Coax Broadband  
10  
Sliding  
Coax  
Coax  
Sliding  
Thru  
11 Delay/  
thru  
12  
Load  
Fixed  
0
50  
50  
0
0
0
2.001  
999  
Coax  
Lowband  
13 Delay/  
thru  
134.82  
2.2  
Coax f-f adapter  
14 Delay/  
thru  
196  
50  
2.2  
0
999  
Coax m-m  
adapter  
15  
16  
17  
18  
19  
20  
21  
a. Ensure system Z0 of network analyzer is set to this value.  
b. Open, short, load, delay/thru, or arbitrary impedance.  
c. Load or arbitrary impedance only.  
d. For waveguide, the lower frequency is the same as FCO.  
e. Typical values only. Disk values may be different.  
f. Standard labels which specify ex, (m) or (f), refer to the sex of the test port connector.  
A-12  
85054B  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Standard Definitions  
Nominal Standard Definitions  
Ta b le A-8  
Sta n d a r d De fin ition s for t h e 872x Se r ie s Net w or k An a lyze r  
System Z0a = 50.0 Ω  
Calibration Kit Label: [N 50Ω]  
F r eq u en cy  
in GH zd  
Sta n d a r d b  
Offset  
1
2
3
4
Short  
Open  
Load  
0
0
0
0
27.99  
50  
800M  
800M  
800M  
800M  
0
999  
999  
999  
999  
Coax  
Coax  
Short (m)  
Open (m)  
88.31  
1667.2 146.61  
9.75  
22.905 50  
0
0
0
Fixed  
0
0
50  
50  
Coax Broadband  
Delay/  
thru  
Coax  
Thru  
5
6
7
8
Load  
Load  
Short  
Open  
Sliding  
Fixed  
0
0
50  
50  
800M 1.999  
999  
2.001  
999  
Coax  
Coax  
Coax  
Coax  
Sliding  
Lowband  
Short (f)  
Open (f)  
800M  
800M  
800M  
0
0
0
63.078 50  
57.993 50  
88.31  
1667.2 146.61  
9.75  
999  
a. Ensure system Z0 of network analyzer is set to this value.  
b. Open, short, load, delay/thru, or arbitrary impedance.  
c. Load or arbitrary impedance only.  
d. For waveguide, the lower frequency is the same as FCO.  
85054B  
A-13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Standard Definitions  
Nominal Standard Definitions  
Ta b le A-9  
St a n d a r d Defin it ion s for th e P NA Se r ie s Net w or k An a lyze r  
System Z0a = 50.0 Ω  
Calibration Kit Label: N 50Ω  
F r eq u e n cy  
in GH zc  
Sta n d a r d b  
Offset  
Short (m)d  
1
2
Short  
Open  
0.1315  
606.21 68.405 2.0206  
144.62 2.2258  
27.990 50 1.3651  
0
999  
999  
Coax  
Coax  
104.13 1943.4  
22.905 50  
0.93  
0
0
Open (m)  
3
Broadband  
load  
Fixed  
0
50  
999  
Coax Broadband  
load  
4
5
Thru  
Thru  
0
0
50  
50  
0
0
999  
999  
Coax  
Coax  
Thru  
Sliding  
load  
Sliding  
Sliding  
load  
6
Lowband  
load  
Fixed  
0
50  
0
999  
Coax  
Lowband  
load  
7
8
Short  
Open  
0.7536  
89.939  
459.88 52.429 1.5846  
2536.8 264.99 13.4  
63.078 50 1.1273  
57.998 50 0.93  
0
0
999  
999  
Coax  
Coax  
Short (f)  
Open (f)  
a. Ensure system Z0 of network analyzer is set to this value.  
b. Open, short, load, delay/thru, or arbitrary impedance.  
c. For waveguide, the lower frequency is the same as FCO.  
d. Standard labels which specify ex, (m) or (f), refer to the sex of the test port connector.  
A-14  
85054B  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Standard Definitions  
Nominal Standard Definitions  
Bla n k For m s  
The standard definitions may be changed to meet your specific requirements. Table A-10  
through A-12 are provided to record the modified standard definitions.  
Ta b le A-10  
Bla n k For m for t h e 8510 Net w or k An a lyze r  
System Z0a = _________________  
Calibration Kit Label: ________________________  
File Number: _________________________________  
Disk File Name: ______________________________  
F r eq u e n cy  
St a n d a r d b  
in GH zd  
Offset  
1
2
3
4
5
6
7
8
9
10  
11  
12  
13  
14  
15  
16  
17  
18  
19  
20  
21  
a. Ensure system Z0 of network analyzer is set to this value.  
b. Open, short, load, delay/thru, or arbitrary impedance.  
c. Load or arbitrary impedance only.  
d. For waveguide, the lower frequency is the same as FCO.  
85054B  
A-15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Standard Definitions  
Nominal Standard Definitions  
Ta b le A-11 Bla n k For m for th e 872x Ser ies of Net w or k An a lyzer s  
System Z0a = 50.0 Ω  
Calibration Kit Label:________________________  
Fr equ en cy  
Offset  
Sta n d a r d b  
in GH zd  
1
2
3
4
5
6
7
8
a. Ensure system Z0 of network analyzer is set to this value.  
b. Open, short, load, delay/thru, or arbitrary impedance.  
c. Load or arbitrary impedance only.  
d. For waveguide, the lower frequency is the same as FCO.  
A-16  
85054B  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Standard Definitions  
Nominal Standard Definitions  
Ta b le A-12 Bla n k For m for t h e P NA Ser ies of Net w or k An a lyzer s  
System Z0a = _______________  
Calibration Kit Label: ________________________  
F r eq u en cy  
Sta n d a r d b  
in GH zc  
Offset  
1
2
3
4
5
6
7
8
a. Ensure system Z0 of network analyzer is set to this value.  
b. Open, short, load, delay/thru, or arbitrary impedance.  
c. For waveguide, the lower frequency is the same as FCO.  
85054B  
A-17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Standard Definitions  
Nominal Standard Definitions  
A-18  
85054B  
Download from Www.Somanuals.com. All Manuals Search And Download.  
In d ex  
A
part numbers, 6-6  
compressed air or nitrogen, 3-4  
conductor  
mating plane, 2-3  
data, recertification, 4-2  
defective connectors, 3-3  
definitions  
adapters, 1-3  
part numbers, 6-6  
Agilent Technologies  
contacting, 5-3  
alcohol  
standard, A-11  
deviation from nominal phase, 2-5  
conductor  
device  
protrusion, 2-3  
conductor  
isopropyl  
as cleaning solvent, 3-4  
recession, 2-3  
mating plane, 2-3  
connecting, 3-16  
disconnecting, 3-19  
handling, 3-22  
connections, 3-2, 3-16, 3-19  
cautions in making, 3-16  
disconnecting, 3-19  
ESD concerns, 3-16  
final, 3-16  
altitude, 2-2  
assistance  
contacting agilent, 5-3  
maintenance, 1-8  
part numbers, 6-6  
pin depth, 2-3  
B
how to make, 3-16  
preliminary, 3-16  
sliding load, 3-18  
undoing, 3-19  
blank forms  
storage, 3-22  
standard class assignments, A-7  
temperature, 2-2  
visual inspection, 3-3  
devices  
standard definitions, A-15  
using a torque wrench, 3-16  
connector  
C
how Agilent verifies, 4-2  
cleaning, 3-4  
dial  
cal kit  
cleaning supplies, 6-6  
connector gage, 3-7  
dimensions  
device  
center conductor, 2-3  
outer conductor, 2-3  
disconnections, 3-19  
documentation warranty, ii  
contents, 1-2  
damage, 3-3  
history, 1-6  
female, 3-4, 3-10  
overview, 1-2  
gage  
calibration  
dial, 3-7  
bad, 5-2  
gaging, 3-6  
certificate of, 4-2  
when to do, 3-7  
constants, 1-3  
permanently stored, 1-3  
life, 3-3  
male, 3-8  
standards, 2-5  
E
mating plane, 3-5  
temperature, 2-2  
calibration constants  
entering, 1-3  
electrical specifications, 2-5  
electrostatic discharge, 3-2  
supplies  
precision slotless (female), 1-7,  
sex,31--48  
permanently stored, 1-3  
calibration definitions, 1-3  
calibration kit  
part numbers, 6-6  
threads, 3-4  
when making connections, 3-16  
environmental requirements, 2-2  
equipment  
visual inspection, 3-3  
wear, 3-3  
contents, 1-2  
connector gage  
handling, 3-8, 3-10  
master, 3-8, 3-10  
reading, 3-7  
history, 1-6  
required  
overview, 1-2  
but not supplied, 6-2, 6-6  
calibration label  
part number, 6-6  
calibration report, 4-2  
calibration sticker, 4-3  
center conductor  
protrusion, 3-7  
recession, 3-7  
supplied, 1-2, 6-2  
equipment required, 1-3  
zeroing, 3-8, 3-10  
connector gage accuracy, 3-6  
constants  
ESD, 3-2  
precautions, 3-2, 3-4  
supplies  
calibration, 1-3  
part numbers, 6-6  
entering, 1-3  
when making connections, 3-16  
permanently stored, 1-3  
contacting Agilent Technologies,  
certificate of calibration, 4-2  
certification  
F
specifications, 2-5  
frequency  
contents  
incomplete  
characteristics  
specifications, 2-5  
mechanical, 2-3  
supplemental, 2-3  
class assignments  
blank forms, A-7  
standard, A-3  
frequency range, 2-5  
what to do, 1-4  
kit, 1-2  
G
D
gage  
cleaning connectors, 3-4  
cleaning supplies, 1-3  
ordering, 6-6  
connector, 1-3  
damage  
dial, 3-7  
to connectors, 3-3  
handling, 3-8, 3-10  
damaged connectors, 3-3  
85054B  
Index-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
In d ex  
master, 3-8, 3-10, 6-6  
mating plane inspection, 3-3  
mating plane surfaces, 3-5  
mechanical characteristics, 2-3  
mechanical integrity, 2-3  
regulations  
reading, 3-7  
environmental, 3-5  
replaceable parts, 6-2  
requirements  
zeroing, 3-8, 3-10  
gage master  
part numbers, 6-6  
environmental, 2-2  
return kit or device to Agilent,  
using, 3-8, 3-10  
N
gaging  
nitrogen, 3-4  
female connectors, 3-10  
male connectors, 3-8  
procedures, 3-8  
return loss  
nominal standard definitions,  
specifications, 2-5  
numbers  
serial, 1-5  
sliding loads, 3-12  
gaging connectors, 3-6  
when to do, 3-7  
S
serial numbers, 1-5  
devices, 1-5  
O
recording, 1-5  
observed limits  
pin depth, 2-4  
H
service, 5-2, 5-3  
service tag, 4-3, 5-3  
set marks  
handling, 3-22  
humidity, 2-2  
offset opens and shorts, 1-2  
open-end wrench, 3-19  
part number, 6-6  
opens, 1-2  
sliding load, 3-20  
sex  
I
connector, 1-8  
impedance  
ordering parts, 6-2  
shorts, 1-2  
system, A-11  
sliding load, 1-3  
calibration, 3-20  
connecting, 3-18  
gaging, 3-12  
setting, A-11  
P
incoming inspection, 1-4  
information, troubleshooting, 5-3  
inspection  
part numbers, 6-2  
parts  
ordering, 6-2  
pin depth  
damage, 3-3  
replaceable, 6-2  
adjusting, 3-13  
set marks, 3-20  
sliding ring, 3-20  
using, 3-20  
defects, 3-3  
required but not supplied, 6-2  
performance verification, 4-2  
fail, 5-2  
incoming, 1-4  
mating plane, 3-3  
visual, 3-3  
permanently stored calibration  
definitions, 1-3  
specifications, 2-2  
isopropyl alcohol  
as cleaning solvent, 3-4  
altitude, 2-2  
certification, 2-5  
pin depth, 2-3  
adjusting sliding load, 3-13  
electrical, 2-5  
K
definition of, 2-3  
environmental, 2-2  
frequency, 2-5  
humidity, 2-2  
kit  
effect on electrical  
specifications, 2-4  
importance of, 2-4  
observed limits, 2-4  
typical values, 2-4  
precision slotless connectors, 1-7,  
contents, 1-2  
overview, 1-2  
pin depth, 2-4  
return loss, 2-5  
temperature, 2-2  
verifying, 4-2  
L
label  
standard class assignments, A-3  
blank forms, A-7  
standard definitions, A-11, A-15  
blank forms, A-15  
nominal, A-11  
standards  
calibration, 4-3  
part number, 6-6  
limits  
pin depth, 2-4  
preventive maintenance, 1-8  
protrusion  
center conductor, 3-7  
conductor, 2-3  
M
R
calibration, 2-5  
NIST, 2-5, 4-3  
static discharge, 3-2  
sticker  
maintenance, 3-2  
reading connector gage, 3-7  
recertification, 4-3  
how often?, 4-3  
preventive, 1-8  
maintenance of devices, 1-8  
making connections, 3-16  
manual  
calibration, 4-3  
interval, 4-3  
storage, 3-22  
whats included, 4-3  
where to send your kit, 4-3  
recession  
part number, 6-6  
altitude, 2-2  
mating plane  
humidity, 2-2  
conductor, 2-3  
temperature, 2-2  
center conductor, 3-7  
connector, 3-5  
supplemental characteristics, 2-3  
conductor, 2-3  
Index-2  
85054B  
Download from Www.Somanuals.com. All Manuals Search And Download.  
In d ex  
supplies  
cleaning, 1-3  
system impedance, A-11  
T
tag  
service, 4-3, 5-3  
temperature  
calibration, 2-2  
device, 2-2  
error-corrected temperature  
range, 2-2  
measurement, 2-2  
operating range, 2-2  
verification and measurement,  
test data, 4-2  
threads  
connector, 3-4  
torque wrench, 1-3, 3-16  
part number, 6-6  
specifications, 3-16  
traceability, 4-2  
troubleshooting, 5-2  
V
verification  
performance, 4-2  
temperature, 2-2  
visual inspection, 3-3  
W
warranty, documentation , ii  
wrench  
open-end, 1-3, 3-19  
part number, 6-6  
undoing connections, 3-19  
torque, 1-3  
part number, 6-6  
wrenches  
part numbers, 6-6  
Z
zeroing  
connector gage, 3-8, 3-10  
zeroing connector gage, 3-8,  
85054B  
Index-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  

3Com Gas Grill Equity I+ User Manual
Aga Ranges Oven U110054 03A User Manual
Air King Ventilation Hood ASLC90 User Manual
Alliance Laundry Systems Washer PHM1263R User Manual
Allied Telesis Network Card AT A35 SX SC User Manual
Alvin Mobility Aid DC207B User Manual
American Audio CD Player DCD PRO240 User Manual
Asus Computer Hardware 90MB0DT0M0AAY0 User Manual
BB Electronics Stereo Amplifier FBDA User Manual
Behringer Music Pedal DM100 User Manual