Agilent Technologies Power Supply 66lxxA User Manual

Programming Guide  
Agilent Technologies  
Series 66lxxA  
MPS Power Modules  
Agilent Part No. 5959-3362  
Microfiche Part No. 5959-3363  
Printed in USA: September 1997  
Update April 2000  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Contents  
1.  
2.  
Introduction  
About this Guide ..........................................................................................................................................7  
Documentation Summary.............................................................................................................................7  
External References......................................................................................................................................7  
VXIPlug&Play Power Products Instrument Drivers....................................................................................8  
Introduction to Programming  
GPIB Capabilities of the Power Module ......................................................................................................9  
Module GPIB Address .................................................................................................................................9  
Introduction to SCPI.....................................................................................................................................9  
Conventions...............................................................................................................................................9  
Types of SCPI Messages.........................................................................................................................10  
Types of SCPI Commands ......................................................................................................................10  
Structure of a SCPI Message...................................................................................................................10  
The Message Unit.................................................................................................................................10  
Combining Message Units....................................................................................................................10  
Parts of a SCPI Message .........................................................................................................................11  
Headers.................................................................................................................................................11  
Query Indicator..........................................................................................................12  
Message Unit Separator........................................................................................................................12  
Root Specifier.......................................................................................................................................12  
Message Terminator .............................................................................................................................12  
Traversing the Command Tree................................................................................................................13  
Active Header Path...............................................................................................................................13  
The Effect of Optional Headers............................................................................................................13  
Moving Among Subsystems.................................................................................................................14  
Including Common Commands...............................................................................................................14  
SCPI Data Formats.....................................................................................................................................14  
Numerical Data........................................................................................................................................14  
Boolean Data...........................................................................................................................................15  
String Data...............................................................................................................................................15  
Character Data.........................................................................................................................................15  
System Considerations ...............................................................................................................................16  
Assigning the Address in Programs.........................................................................................................16  
DOS Drivers............................................................................................................................................16  
Types of Drivers...................................................................................................................................16  
Agilent 82335A Driver.........................................................................................................................16  
National Instruments GPIB Driver .......................................................................................................16  
Error Handling......................................................................................................................................17  
Agilent BASIC for Series 300.................................................................................................................17  
Translation Among Languages...................................................................................................................17  
General Setup Information for GWBASIC .............................................................................................17  
Using the Agilent 82335A/82990A/16062B GPIB Command Library ................................................17  
Using the National Instruments GPIB Interface ...................................................................................18  
General Setup Information for Microsoft C ............................................................................................18  
Using the Agilent 82335A/82990A/16062B GPIB Command Library ................................................18  
Using the National Instruments GPIB Interface ...................................................................................18  
Sending Commands to and Receiving Data from the Module.................................................................19  
3.  
Language Dictionary  
Introduction ................................................................................................................................................23  
Parameters..................................................................................................................................................23  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
Related Commands.....................................................................................................................................23  
Order of Presentation .................................................................................................................................23  
Common Commands..................................................................................................................................23  
Subsystem Commands................................................................................................................................23  
Description of Common Commands ..........................................................................................................24  
*CLS........................................................................................................................................................24  
*ESE........................................................................................................................................................25  
*ESR?......................................................................................................................................................25  
*IDN?......................................................................................................................................................26  
*OPC.......................................................................................................................................................26  
*OPC? .....................................................................................................................................................26  
*OPT? .....................................................................................................................................................27  
*PSC........................................................................................................................................................27  
*RCL.......................................................................................................................................................28  
*RST .......................................................................................................................................................28  
*SAV.......................................................................................................................................................29  
*SRE .......................................................................................................................................................29  
*STB?......................................................................................................................................................30  
*TRG.......................................................................................................................................................31  
*TST?......................................................................................................................................................31  
*WAI.......................................................................................................................................................31  
Description of Subsystem Commands........................................................................................................31  
ABOR......................................................................................................................................................31  
Calibration Subsystem.............................................................................................................................32  
CAL:AUTO..........................................................................................................................................32  
CAL:CURR..........................................................................................................................................33  
CAL:CURR:LEV .................................................................................................................................33  
CAL:PASS ...........................................................................................................................................33  
CAL:SAV.............................................................................................................................................34  
CAL:STAT...........................................................................................................................................34  
CAL:VOLT ..........................................................................................................................................34  
CAL:VOLT:LEV .................................................................................................................................34  
CAL:VOLT:PROT...............................................................................................................................35  
Current Subsystem...................................................................................................................................35  
CURR...................................................................................................................................................35  
CURR:MODE ......................................................................................................................................35  
CURR:PROT:STAT.............................................................................................................................36  
CURR:TRIG.........................................................................................................................................36  
DISP:STAT ..........................................................................................................................................36  
INIT......................................................................................................................................................36  
INIT:CONT..........................................................................................................................................36  
List Subsystem.........................................................................................................................................37  
LIST:COUN.........................................................................................................................................37  
LIST:CURR .........................................................................................................................................37  
LIST:CURR:POIN? .............................................................................................................................38  
LIST:DWEL.........................................................................................................................................38  
LIST:DWEL:POIN?.............................................................................................................................38  
LIST:STEP...........................................................................................................................................38  
LIST:VOLT..........................................................................................................................................39  
LIST:VOLT:POIN? .............................................................................................................................39  
MEAS:CURR? .....................................................................................................................................39  
MEAS:VOLT? .....................................................................................................................................39  
Output Subsystem....................................................................................................................................39  
OUTP ...................................................................................................................................................39  
OUTP:DFI............................................................................................................................................40  
4
Download from Www.Somanuals.com. All Manuals Search And Download.  
OUTP:DFI:LINK .................................................................................................................................40  
OUTP:DFI:SOUR ................................................................................................................................40  
OUTP:PROT:CLE ...............................................................................................................................40  
OUTP:PROT:DEL ...............................................................................................................................41  
OUTP:REL...........................................................................................................................................41  
OUTP:REL:POL ..................................................................................................................................41  
OUTP:TTLT ........................................................................................................................................42  
OUTP:TTLT:LINK..............................................................................................................................42  
OUTP:TTLT:SOUR.............................................................................................................................42  
Status Subsystem.....................................................................................................................................42  
STAT:OPER?.......................................................................................................................................43  
STAT:OPER:COND? ..........................................................................................................................43  
STAT:OPER:ENAB.............................................................................................................................43  
STAT:OPER:NTR ...............................................................................................................................44  
STAT:OPER:PTR ................................................................................................................................44  
STAT:PRES .........................................................................................................................................44  
STAT:QUES? ......................................................................................................................................45  
STAT:QUES:COND? ..........................................................................................................................45  
STAT:QUES:ENAB ............................................................................................................................45  
STAT:QUES:NTR ...............................................................................................................................45  
STAT:QUES:PTR................................................................................................................................45  
SYST:ERR? .........................................................................................................................................46  
SYST:VERS?.......................................................................................................................................46  
Trigger Subsystem...................................................................................................................................46  
TRIG ....................................................................................................................................................46  
TRIG:DEL............................................................................................................................................47  
TRIG:LINK..........................................................................................................................................47  
TRIG:SOUR.........................................................................................................................................47  
Voltage Subsystem..................................................................................................................................48  
VOLT ...................................................................................................................................................48  
VOLT:MODE ......................................................................................................................................48  
VOLT:PROT........................................................................................................................................48  
VOLT:SENS:SOUR?...........................................................................................................................49  
VOLT:TRIG.........................................................................................................................................49  
Link Parameter List.................................................................................................................................50  
Power Module Programming Parameters................................................................................................50  
4.  
Status Reporting  
Power Module Status Structure ..................................................................................................................51  
Status Register Bit Configuration...............................................................................................................51  
Operation Status Group..............................................................................................................................51  
Register Functions...................................................................................................................................51  
Register Commands.................................................................................................................................51  
Questionable Status Group.........................................................................................................................52  
Register Functions...................................................................................................................................52  
Register Commands.................................................................................................................................52  
Standard Event Status Group......................................................................................................................53  
Register Functions...................................................................................................................................53  
Register Commands.................................................................................................................................53  
Status Byte Register ...................................................................................................................................54  
The RQS Bit............................................................................................................................................54  
The MSS Bit............................................................................................................................................54  
Determining the Cause of a Service Interrupt..........................................................................................54  
Output Queue .............................................................................................................................................54  
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
Location of Event Handles .........................................................................................................................54  
Initial Conditions at Power On ...................................................................................................................55  
Status Registers .......................................................................................................................................55  
The PON Bit............................................................................................................................................55  
Examples....................................................................................................................................................55  
Servicing an Operation Status Event .......................................................................................................55  
Adding More Operation Events...............................................................................................................56  
Servicing Questionable Status Events .....................................................................................................56  
Monitoring Both Phases of a Status Transition .......................................................................................56  
5.  
Synchronizing Power Module Output Changes  
Introduction ................................................................................................................................................57  
Trigger Subsystem......................................................................................................................................57  
Model of Fixed-Mode Trigger Operation................................................................................................57  
Idle State...............................................................................................................................................58  
Initiated State........................................................................................................................................58  
Delaying State ......................................................................................................................................58  
Output Change State.............................................................................................................................59  
Model of List-Mode Trigger Operation...................................................................................................59  
Output Change State.............................................................................................................................59  
Dwelling State ......................................................................................................................................59  
The INIT:CONT Function.......................................................................................................................59  
Trigger Status and Event Signals.............................................................................................................59  
Trigger In and Trigger Out......................................................................................................................60  
List Subsystem............................................................................................................................................61  
Basic Steps of List Sequencing ...............................................................................................................61  
Programming the List Output Levels.......................................................................................................61  
Programming List Intervals.....................................................................................................................61  
Automatically Repeating a List ...............................................................................................................62  
Triggering a List......................................................................................................................................62  
Dwell-Paced Lists.................................................................................................................................62  
Trigger-Paced Lists ..............................................................................................................................62  
DFI (Discrete Fault Indicator) Subsystem..................................................................................................64  
RI (Remote Inhibit) Subsystem ..................................................................................................................64  
SCPI Command Completion ......................................................................................................................64  
6.  
Error Messages  
Power Module Hardware Error Messages..................................................................................................65  
System Error Messages ..............................................................................................................................65  
A.  
B.  
SCPI Conformance Information............................................................................................................67  
Application Programs..............................................................................................................................69  
Index ..........................................................................................................................................................111  
6
Download from Www.Somanuals.com. All Manuals Search And Download.  
1
Introduction  
About This Guide  
You will find the following information in the rest of this guide:  
Chapter 2  
Chapter 3  
Chapter 4  
Chapter 5  
Chapter 6  
Appendix A  
Appendix B  
Introduction to SCPI messages structure, syntax, and data formats. Examples of SCPI programs.  
Dictionary of SCPI commands. Table of module programming parameters.  
Description of the status registers.  
Description of synchronizing outputs with triggers and lists.  
Error messages.  
SCPI conformance information.  
Application programs that illustrate features of the power module.  
Note  
Instructions for the Agilent 60001A MPS Keyboard are in the User’s Guide supplied with each module.  
Documentation Summary  
The following related documents shipped with the system have information helpful to programming the power module:  
Mainframe User’s Guide. Information on the GPIB address switch, trigger connections, fault (FLT) and  
remote inhibit (INH) connections.  
Module User’s Guide. Includes specifications and supplemental characteristics, use of the module  
configuration switch, device related error messages, calibration procedures and use of the MPS keyboard.  
External References  
SCPI References  
The following documents will assist you with programming in SCPI:  
Beginner’s Guide to SCPI. Part No. H2325-90001. Highly recommended for anyone who has not had previous  
experience programming with SCPI.  
Tutorial Description of the GPIB . Part No. 5952-0156. Highly recommended for those not familiar with the  
IEEE 488.1 and 488.2 standards.  
To obtain a copy of the above documents, contact your local Agilent Technologies Sales and Support Office.  
GPIB References  
The most important GPIB documents are your controller programming manuals - GW BASIC, GPIB Command Library for  
MS DOS, etc. Refer to these for all non-SCPI commands (for example: Local Lockout).  
The following are two formal documents concerning the GPIB interface:  
2 ANSI/IEEE Std. 488.1-1987 IEEE Standard Digital Interface for Programmable Instrumentation . Defines  
the technical details of the GPIB interface. While much of the information is beyond the need of most  
programmers, it can serve to clarify terms used in this guide and in related documents.  
2 ANSI/IEEE Std. 488.2-1987 IEEE Standard Codes, Formats, Protocols, and Common Commands .  
Recommended as a reference only if you intend to do fairly sophisticated programming. Helpful for finding  
precise definitions of certain types of SCPI message formats, data types, or common commands.  
The above two documents are available from the IEEE (Institute of Electrical and Electronics Engineers), 345 East 47th  
Street, New York, NY 10017, USA.  
Introduction  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
VXIplug&play Power Products Instrument Drivers  
VXIplug&play instrument drivers for Microsoft Windows 95 and Windows NT are now available on the Web at  
http://www.agilent.com/find/drivers. These instrument drivers provide a high-level programming interface to your Agilent  
Technologies electronic load. VXIplug&play instrument drivers are an alternative to programming your instrument with  
SCPI command strings. Because the instrument driver’s function calls work together on top of the VISA I/O library, a  
single instrument driver can be used with multiple application environments.  
Supported Applications  
ñ
ñ
ñ
ñ
ñ
ñ
Agilent VEE  
Microsoft Visual BASIC  
Microsoft Visual C/C++  
Borland C/C++  
National Instruments LabVIEW  
National Instruments LabWindows/CVI  
System Requirements  
The VXIplug&play instrument driver complies with the following:  
ñ
ñ
ñ
ñ
Microsoft Windows 95  
Microsoft Windows NT 4.0  
HP VISA revision F.01.02  
National Instruments VISA 1.1  
Downloading and Installing the Driver  
NOTE:  
Before installing the VXIplug&play instrument driver, make sure that you have one of the supported  
applications installed and running on your computer.  
1. Access Agilent Technologies Web site at http://www.agilent.com/find/drivers.  
2. Select the instrument for which you need the driver.  
3. Click on the driver, either Windows 95 or Windows NT, and download the executable file to your PC.  
4. Locate the file that you downloaded from the Web. From the Start menu select Run <path>:\agxxxx.exe - where  
<path> is the directory path where the file is located, and agxxxx is the instrument driver that you downloaded .  
5. Follow the directions on the screen to install the software. The default installation selections will work in most  
cases. The readme.txt file contains product updates or corrections that are not documented in the on-line help. If  
you decide to install this file, use any text editor to open and read it.  
6. To use the VXIplug&play instrument driver, follow the directions in the VXIplug&play online help for your  
specific driver under “Introduction to Programming”.  
Accessing Online Help  
A comprehensive online programming reference is provided with the driver. It describes how to get started using the  
instrument driver with Agilent VEE, LabVIEW, and LabWindows. It includes complete descriptions of all function calls  
as well as example programs in C/C++ and Visual BASIC.  
ñ
To access the online help when you have chosen the default Vxipnp start folder, click on the Start button and select  
Programs | Vxipnp | Agxxxx Help (32-bit).  
- where Agxxxx is the instrument driver.  
8
Introduction  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2
Introduction To Programming  
GPIB Capabilities Of The Power Module  
All power module functions except for setting the GPIB address are programmable over the GPIB. The IEEE 488.1  
capabilities of the power module are listed in the User’s Guide.  
Module GPIB Address  
The power module operates from a primary GPIB address that is set by a switch on the mainframe. The power module’s  
secondary GPIB address is determined by its slot position within the mainframe. See the mainframe Installation Guide for  
details.  
Introduction To SCPI  
SCPI (Standard Commands for Programmable Instruments) is a programming language for controlling instrument functions  
over the GPIB (IEEE 488) instrument bus. SCPI is layered on top of the hardware-portion of IEEE 488.2. The same SCPI  
commands and parameters control the same functions in different classes of instruments. For example, you would use the  
same DISPlay command to control the power module display state and the display state of a SCPI-compatible multimeter.  
Conventions  
The following conventions are used throughout this chapter:  
Angle brackets < >  
Items within angle brackets are parameter abbreviations. For example, <NR1>  
indicates a specific form of numerical data.  
Vertical bar  
|
Vertical bars separate one of two or more alternative parameters. For example,  
0|OFF indicates that entering either "0" or "OFF" performs the same function.  
Square Brackets [ ]  
Items within square brackets are optional. The representation [SOURce]:LIST  
means that SOURce may be omitted.  
Braces  
{ }  
Braces indicate parameters that may be repeated zero or more times. It is used  
especially for showing arrays. The notation <A>{<,B>} shows that parameter  
"A" must be entered, while parameter "B" may be omitted or may be entered one  
or more times.  
Boldface font  
Boldface font is used to emphasize syntax in command definitions.  
TRIGger:DELay <NRf> shows a command definition.  
Computer font  
Computer font is used to show program lines in text. TRIGger: DELay .5 shows a  
program line.  
Introduction To Programming  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
Types of SCPI Messages  
There are two types of SCPI messages, program and response.  
A program message consists of one or more properly formatted SCPI commands sent from the controller to  
the power module. The message, which may be sent at any time, requests the power module to perform some  
action.  
A response message consists of data in a specific SCPI format sent from the power module to the controller.  
The power module sends the message only when commanded by a special program message called a "query."  
Types of SCPI Commands  
SCPI has two types of commands, common and subsystem. Common commands generally are not related to specific  
operation but to controlling overall power module functions, such as reset, status, and synchronization. All common  
commands consist of a three-letter mnemonic preceded by an asterisk:  
*RST *IDN? *SRE 8  
Subsystem commands perform specific power module functions. They are organized into an inverted tree structure with the  
"root" at the top (see Figure 3-2). Some are single commands while others are grouped within specific subsystems.  
Note  
If you have the optional Agilent 66001A MPS Keyboard, you may want to use it as a quick introduction  
to message structure. See "Appendix A".  
Structure of a SCPI Message  
SCPI messages consist of one or more message units ending in a message terminator. The terminator is not part of the  
syntax, but implicit in the way your programming language indicates the end of a line (such as a newline or end-of-line  
character).  
The Message Unit  
The simplest SCPI command is a single message unit consisting of a command header (or keyword) followed by a message  
terminator.  
ABOR<newline>  
VOLT?<newline>  
The message unit may include a parameter after the header. The parameter usually is numeric, but it can be a string:  
VOLT 20<newline>  
VOLT MAX<newline>  
Combining Message Units  
The following command message is briefly described here, with details in subsequent paragraphs.  
10 Introduction To Programming  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Figure 2-1. Command Message Structure  
The basic parts of the above message are:  
Message Component  
Headers  
Example  
VOLT LEV PROT CURR  
Header Separator  
Data  
The colon in VOLT:LEV  
8.0 8.8  
Data Separator  
Message Units  
Message Unit Separator  
Root Specifier  
Query Indicator  
Message Terminator  
The space in VOLT 8. 0 and PROT 8. 8  
VOLT:LEV 8.0 PROT 8.8 CURR?  
The semicolons in VOLT: LEV 8. 0; and PROT 8. 8;  
The colon in PROT 8. 8; : CURR?  
The question mark in CURR?  
The <NL> (newline) indicator. Terminators are not part of the SCPI syntax.  
Parts of a SCPI Message  
Headers  
Headers are instructions recognized by the power module interface. Headers (which are sometimes known as "keywords")  
may be either in the long form or the short form.  
Long Form  
Short Form  
The header is completely spelled out, such as VOLTAGE, STATUS, and DELAY.  
The header has only the first three or four letters, such as VOLT, STAT, and DEL.  
Short form headers are constructed according to the following rules:  
If the header consists of four or fewer letters, use all the letters. (DFI  
If the header consists of five or more letters and the fourth letter is not a vowel (a,e,i,o,u), use the first four  
letters. (CURRent STATus)  
LIST)  
If the header consists of five or more letters and the fourth letter is a vowel (a,e,i,o,u), use the first three letters.  
(DELay RELay)  
You must follow the above rules when entering headers. Creating an arbitrary form, such as POLAR for POLarity, will result  
in an error.  
Introduction To Programming 11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
The SCPI interface is not sensitive to case. It will recognize any case mixture, such as TRIGGER, Trigger, TRIGger, triGgeR.  
Shortform headers result in faster program execution.  
Note  
Header Convention. In this manual, headers are emphasized with boldface type. The proper short form is shown in  
upper-case letters, such as DELay.  
Header Separator. If a command has more than one header, you must separate them with a colon (VOLT: PROT  
OUTPut:RELay:POLarity).  
Optional Headers. The use of some headers is optional. Optional headers are shown in brackets, such as  
OUTPut[: STATe] ON. However, if you combine two or more message units into a compound message, you may need to  
enter the optional header. This is explained under "Traversing the Command Tree."  
Note  
The optional Agilent 66001A MPS Keyboard does not display optional headers.  
Query Indicator  
Following a header with a question mark turns it into a query (VOLT?, VOLT:PROT?). If a query contains a parameter, place  
the query indicator at the end of the last header (VOLT: PROT? MAX).  
Message Unit Separator  
When two or more message units are combined into a compound message, separate the units with a semicolon  
(STATus:OPERation?;QUEStionable?). You can combine message units only at the current path of the command tree (see  
"Traversing the Command Tree").  
Root Specifier  
When it precedes the first header of a message unit, the colon becomes the root specifier. It indicates that the parser is at the  
root or top node of the command tree. Note the difference between root specifiers and header separators in the following  
examples:  
OUTP:PROT:DEL .1  
All colons are header separators.  
:OUTP:PROT:DEL .1  
OUTP: PROT: DEL . 1; :VOLT 12.5  
Only the first colon is a root specifier.  
Only the third colon is a root specifier.  
Message Terminator  
A terminator informs SCPI that it has reached the end of a message. Three permitted messages terminators are:  
newline (<NL>), which is ASCII decimal 10 or hex 0A.  
end or identify (<END>)  
both of the above (<NL><END>).  
In the examples of this guide, there is an assumed message terminator at the end of each message. If the terminator needs to  
be shown, it is indicated as <NL> regardless of the actual terminator character.  
12 Introduction To Programming  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Traversing the Command Tree  
Figure 2-2 shows a portion of the subsystem command tree (you can see the complete tree in Figure 3-2). Note the location  
of the ROOT node at the top of the tree. The SCPI interface is at this location when:  
The power module is powered on.  
A device clear (DCL) is sent to the power module.  
The interface encounters a message terminator.  
The interface encounters a root specifier.  
Figure 2-2. Partial Command Tree  
Active Header Path  
In order to properly traverse the command tree, you must understand the concept of the active header path. When the power  
module is turned on (or under any of the other conditions listed above), the active path is at the root. That means the  
interface is ready to accept any command at the root level, such as OUTPUT or STATUS in Figure 2-2. Note that you do not  
have to proceed either command with a colon; there is an implied colon in front of every root-level command.  
If you enter OUTPUT, the active header path moves one colon to the right. The interface is now ready to accept :STATE,  
:PROTECTION, or :RELAY as the next header. Note that you must include the colon, because it is required between headers.  
If you now enter :PROTECTION, the active path again moves one colon to the right. The interface is now ready to accept  
either :CLEAR or :DELAY as the next header.  
If you now enter :CLEAR, you have reached the end of the command string. The active header path remains at :CLEAR. If  
you wished, you could have entered :CLEAR; DELAY 20 and it would be accepted. The entire message would be  
OUTPUT:PROTECTION:CLEAR;DELAY 20. The message terminator after DELAY 20 returns the path to the root.  
The Effect of Optional Headers  
If a command includes optional headers, the interface assumes they are there. For example, if you enter OUTPUT OFF, the  
interface recognizes it as OUTPUT: STATE OFF (see Figure 2-2). This returns the active path to the root (:OUTPUT). But if  
you enter OUTPUT: STATE OFF, then the active path remains at :STATE. This allows you to send OUTPUT: STATE OFF;  
PROTECTION: CLEAR in one message. If you tried to send OUTPUT OFF;PROTECTION:CLEAR, the header path would return  
to :OUTPUT instead of :PROTECTION.  
Introduction To Programming 13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
The optional header SOURCE precedes the current, list, and voltage subsystems (see Figure 3-2). This effectively makes  
:CURRENT, :LIST, and :VOLTAGE root-level commands.  
Note  
The optional Agilent 66001 Keyboard does not display optional headers.  
Moving Among Subsystems  
In order to combine commands from different subsystems, you need to be able to restore the active path to the root. You do  
this with the root specifier (:). For example, you could clear the output protection and check the status of the Operation  
Condition register as follows (see Figure 2-2):  
OUTPUT:PROTECTION:CLEAR  
STATUS:OPERATION:CONDITION?  
By using the root specifier, you could do the same thing in one message:  
OUTPUT:PROTECTION:CLEAR;:STATUS:OPERATION:CONDITION?  
Note  
The SCPI parser traverses the command tree as described in Appendix A of the IEEE 488.2 standard.  
The "Enhanced Tree Walking Implementation" given in that appendix is not implemented in the power  
module.  
The following message shows how to combine commands from different subsystems as well as within the same subsystem  
(see Figure 3-2):  
VOLTAGE:LEVEL 7;PROTECTION 8;:CURRENT:LEVEL 3;MODE LIST  
Note the use of the optional header LEVEL to maintain the correct path within the voltage and current subsystems and the  
use of the root specifier to move between subsystems.  
Including Common Commands  
You can combine common commands with system commands in the same message. Treat the common command as a  
message unit by separating it with the message unit separator. Common commandsdo not affect the active header path;  
you may insert them anywhere in the message.  
VOLT:TRIG 7.5;INIT;*TRG  
OUTP OFF;*RCL 2;OUTP ON  
SCPI Data Formats  
All data programmed to or returned from the power module is ASCII. The data may be numerical or character string.  
Numerical Data  
Table 2-1 and Table 2-2 summarize the numerical formats.  
14 Introduction To Programming  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table 2-1. Numerical Data Formats  
Data Form  
Symbol  
<NR1>  
Talking Formats  
Digits with an implied decimal point assumed at the right of the least-significant  
digit. Examples: 273 0273  
<NR2>  
<NR3>  
Digits with an explicit decimal point. Example: 273. .0273  
Digits with an explicit decimal point and an exponent. Example: 2.73E+2 273.0E-2  
Listening Formats  
<NRf>  
Extended format that includes <NR1>, <NR2> and <NR3>. Examples: 273 273.  
2.73E2  
<NRf+>  
Expanded decimal format that includes <NRf> and MIN MAX. Examples: 273 273.  
2.73E2 MAX. MIN and MAX are the minimum and maximum limit values that are  
implicit in the range specification for the parameter.  
Table 2-2. Suffixes and Multipliers  
Class  
Current  
Amplitude  
Time  
Suffix  
Unit  
Ampere  
Volt  
second  
Unit with Multiplier  
MA (milliampere)  
MV (millivolt)  
A
V
S
MS (millisecond)  
Common Multipliers  
lE3  
1E-3  
1E-6  
K
M
U
kilo  
milli  
micro  
Boolean Data  
Either form 1 | 0 or ON | OFF may be sent with commands. Queries always return 1 or 0.  
OUTPut OFF  
CURRent:PROTection 1  
String Data  
Strings are used for both program (listening) and response (talking) data. String content is limited to the characters required  
for the link command parameters (see "Chapter 3 - Language Dictionary").  
Note  
The IEEE 488.2 format for a string parameter requires that the string be enclosed within either single  
(’ ’) or double (" ") quotes. Be certain that your program statements comply with this requirement.  
Character Data  
Character strings returned by query statements may take either of the following forms, depending on the length of the  
returned string:  
<CRD>  
<AARD>  
Character Response Data. Permits the return of character strings.  
Arbitrary ASCII Response Data. Permits the return of undelimited 7-bit ASCII. This data type has an  
implied message terminator.  
Introduction To Programming 15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
System Considerations  
The remainder of this chapter addresses some system issues concerning programming. These are power module addressing  
and the use of the following types of GPIB system interfaces:  
1.  
2.  
3.  
HP Vectra PC controller with Agilent 82335A GPIB Interface Command Library  
IBM PC controller with National Instruments GPIB-PCII Interface/Handler  
Agilent controller with Agilent BASIC Language System  
Note  
Some specific application programs are given in Appendix B.  
Assigning the GPIB Address in Programs  
The power module address cannot be set remotely. It is determined by the position of the mainframe address switch and the  
position of power module (slot position) within the mainframe. ( See the Mainframe Users Guide for details.)  
The following examples assume that the GPIB select code is 7, the mainframe interface address is 6, and that the power  
module address will be assigned to the variable PM3 (power module in the third mainframe slot).  
1060 ! Power Module installed in Primary Mainframe  
1070 PM3=70602  
1070 ASSIGN @PM3TO 70602  
1080 !  
! Agilent 82335A Interface  
! Agilent BASIC Interface  
1080 ! Power Module installed in Auxiliary Mainframe  
1090 PM=70610  
! Agilent 82335A Interface  
1090 ASSIGN @PM3 TO 70610 ! Agilent BASIC Interface  
For systems using the National Instruments DOS driver, the address is specified in the software configuration program  
(IBCONFIG.EXE) and assigned a symbolic name. The address then is referenced only by this name within the application  
program (see the National Instruments GPIB documentation).  
DOS Drivers  
Types of Drivers  
The Agilent 82335A and National Instruments GPIB are two popular DOS drivers. Each is briefly described here. See the  
software documentation supplied with the driver for more details.  
Agilent 82335A Driver. For GW-BASIC programming, the GPIB library is implemented as a series of subroutine calls.  
To access these subroutines, your application program must include the header file SETUP.BAS, which is part of the DOS  
driver software.  
SETUP.BAS starts at program line 5 and can run up to line 999. Your application programs must begin at line 1000.  
SETUP.BAS has built-in error checking routines that provide a method to check for GPIB errors during program execution.  
You can use the error-trapping code in these routines or write your own code using the same variables as used by  
SETUP.BAS.  
National Instruments GPIB Driver. Your program must include the National Instruments header file DECL.BAS. This  
contains the initialization code for the interface. Prior to running any applications programs, you must set up the interface  
with the configuration program (IBCONF.EXE).  
16 Introduction To Programming  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Your application program will not include the power module symbolic name and GPIB address. These must be specified  
during configuration (when you run IBCONF.EXE). Note that the primary address range is from 0 to 30 but any secondary  
address must be specified in the address range of 96 to 126. The power supply expects a message termination on EOI or  
line feed, so set EOI w/last byte of Write. It is also recommended that you set Disable Auto Serial Polling.  
All function calls return the status word IBSTA%, which contains a bit (ERR) that is set if the call results in an error. When  
ERR is set, an appropriate code is placed in variable IBERR %. Be sure to check IBSTA %, after every function call. If it is  
not equal to zero, branch to an error handler that reads IBERR% to extract the specific error.  
Error Handling  
If there is no error-handling code in your program, undetected errors can cause unpredictable results. This includes  
"hanging up" the controller and forcing you to reset the system. Both of the above DOS drivers have routines for detecting  
program execution errors.  
Important  
Use error detection after every call to a subroutine.  
Agilent BASIC Controllers  
The Agilent BASIC Programming Language provides access to GPIB functions at the operating system level. This makes it  
unnecessary to have the header files required in front of DOS applications programs. Also, you do not have to be concerned  
about controller "hangups" as long as your program includes a timeout statement. Because the power module can be  
programmed to generate SRQ on errors, your program can use an SRQ service routine for decoding detected errors. The  
detectable errors are listed in "Chapter 5 - Error Messages".  
TRANSLATION AMONG LANGUAGES  
This section explains how to translate between Agilent BASIC and several other popular programming environments. For  
explicit information on initializing interface cards or syntax of language, see the documentation that accompanies your  
GPIB interface product.  
General Setup Information for GWBASIC  
Using the Agilent 82335A/82990A/61062B GPIB Command Library  
When CALLs are made to the GPIB Command Library, all parameters are passed as variables.  
The address of a module is a real number, determined in the same manner as in Agilent BASIC. For example,  
the address 70501 means 7 is the select code of the GPIB interface, 05 is the GPIB address of the mainframe, 01 is  
the slot number (secondary address) of the module.  
The module expects each command to be terminated by line feed (character 10) and/or EOI. The default  
configuration of the GPIB Command Library is carriage return + line feed for end-of-line termination and EOI at  
the end of a line. Therefore, the defaults are correct for use with the module.  
The GPIB Command Library supports strings, numeric and array data formats. However, multiple data types  
cannot be sent in a single command. To send both string and numeric data in one command, convert all numeric  
data to strings, concatenate with the string data and send the combined string to the module. To read multiple data  
types, read the data into a string, and then manipulate the string by converting each piece into the appropriate data  
format.  
Error handling is accomplished by checking the variable PCIB.ERR. If it is nonzero, an error has occurred.  
See the command library documentation for trapping and interpreting this error variable.  
Introduction To Programming 17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Using the National Instruments GPIB Interface  
When CALLs are made to the GPIB driver, all parameters are passed as variables.  
The module is identified as a device in two ways. First, the GPIB.COM driver is modified to include the  
module. Use the mainframe address as the primary bus address and the slot address as the secondary address. The  
driver requires secondary address 0 (which is for slot 0) to be entered as 96, secondary address 1 to be entered as  
97, etc.  
It is recommended that you disable auto serial poll in the GPIB.COM driver.  
The module expects each command to be terminated by a line feed (character 10) and/or EOI. Configure the  
GPIB.COM driver to terminate all reads and writes with EOI.  
The GPIB driver does all communication via strings. To send numeric data, number to-string conversion must  
be performed before the IBWRT( ). To read numeric data, string-to-number conversion must be performed after  
each IBRD( ).  
Error handling is accomplished by checking the variable IBSTA%. If it is less than zero, an error has occurred.  
See the GPIB interface documentation for trapping and interpreting this error variable.  
General Setup Information for Microsoft C  
Using the Agilent 82335A/82990A/61062B GPIB Command Library  
The address of a module is of type long and is determined the same as with Agilent BASIC. For example, the  
address 70501L means 7 is the select code of the GPIB interface, 05 is the GPIB address of the mainframe, 01 is  
the slot number (secondary address) of the module.  
The module expects each command to be terminated by a line feed (character 10) and/or EOI. The default  
configuration of the GPIB Command Library is carriage return+line feed for end-of-line termination and EOI at the  
end of a line. Therefore, the defaults are correct for use with the module.  
The GPIB Command Library supports strings, numeric and array data formats. However, multiple data types  
cannot be sent in a single command. To send both string and numeric data in one command, convert all numeric  
data to strings, concatenate with the string data and send the combined string to the module. To read multiple data  
types, read the data into a string, and then manipulate the string by converting each piece into the appropriate data  
format.  
Each command library call returns an int. If the value is zero, no error has occurred. Error handling is  
accomplished by checking the return value. See the command library documentation for interpretation of this error  
value.  
Using the National Instruments GPIB Interface  
The module is identified as a device in two ways. First, the GPIB.COM driver is modified to include the  
module. Use the mainframe address as the primary bus address. Use the slot address as the secondary address.  
The driver requires that secondary address 0 (which is for slot 0) be entered as 96, secondary address 1 be entered  
as 97, etc.  
It is recommended that you disable the auto serial poll in the GPIB.COM driver.  
The module expects each command to be terminated by either a line feed (character 10) and/or EOI.  
Configure the GPIB.COM driver to terminate all reads and writes with EOI.  
The GPIB driver does all communication via strings. To send numeric data, number to-string conversion  
must be performed before the ibwrt( ). To read numeric data, string-to-number conversion must be performed after  
each ibrd( ).  
Error handling is accomplished by checking the variable IBSTA%. If bit 15 is set, an error has occurred. See  
the GPIB interface documentation for the interpretation of this error variable.  
18 Introduction To Programming  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Sending Commands to and Receiving Data from the Module  
Sending the Command “VOLT 5”  
********************************************A**gilent BASIC ************************************************  
2100 OUTPUT 70501;"VOLT 5" ! where 70501 means 7 is the select code of the GPIB interface 05 is the  
2110  
2120  
!
!
GPIB address of the mainframe, 01 is the slot number (secondary address)  
of the module  
************************* GWBASICA(gilent 82335A/82990A/61062B GPIB Command Library) ************************  
2100 MODULE.ADDRESS=70501  
2110 COMMAND$ - "VOLT 5"  
2120 L - LENGTH(COMMAND$)  
2130 CALL IOOUTPUTS(MODULE.ADDRESS,COMMAND$,L)  
2140 IF PCIB.ERR&<>O THEN ERROR PCIB.BASERR’  
! ERROR TRAP  
********************************** GWBASIC (National Instruments GPIB Interface) ********************************  
2100 COMMAND$ = "VOLT 5"  
2110 CALL IBWRT(MODULE.ADDRESS%, COMMAND$)  
2120 IF IBSTA% &< 0 GOTO 5000  
2130  
! TRAP ERROR WITH ERROR HANDLER  
AT LINE 5000  
!
Sending the Command “VOLT 5” in BASIC  
*********************** Microsoft CA(gilent 82335A/82990A/61062B HPIB Command Library) *************************  
/* Assumes that you have an error handler routine, called ‘error_handler’ that accepts a float. The error handler in then passed the float  
that is returned from each call to the library. */  
*include &<stdio.h>  
*include &<chpib.h>  
*include &<cfunc.h>  
*define module_address 70501L  
char *cmd;  
cmd = "VOLT 5”;  
error = iooutputs(MODULE_ADDRESS, cmd, strlen(cmd));  
error_handler(error);  
****************************** Microsoft C (National Instruments GPIB Interface) ************************************  
/* Assumes that you have an error handler routine, called ‘error-handler’. The error handler is then passed the float that is returned from  
each call to the library. */  
*include &<stdio.h>  
*include &<decl.h>  
*define ERR (l&<&<15) /* Error is detected as bit i5 of ibsta */  
int module-address; /* Device is configured in the GPIB.COM handler. Use ibfind() to assign a value to module-address. */  
char *cmd;  
cmd - "VOLT 5”  
ibwrt(MODULE_ADDRESS, cmd, strlen(cmd));  
if (ibsta & ERR)  
error_handler();  
Sending the Command "VOLT 5” in C  
Introduction To Programming 19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Receiving Data from the Module  
The following screens show how to enter data from the module with various interfaces.  
******************************************* Agilent BASIC ******************************************  
2100 ENTER 70501; MEASUREMENT ! where 70501 means 7 is the select code of the GPIB interface,  
2110 !  
2120 !  
2130 !  
05 is the GPIB address of the mainframe, 01 is the slot number  
(secondary address) of the module, MEASUREMENT is a real number  
sent by the module  
**************** GWBASIC (Agilent 82335A/82990A/61062B GPIB Command Library) ***********************  
2100 MODULE.ADDRESS=70501  
2110 CALL IOENTER(MODULE.ADDRESS,MEASUREMENT)  
2120 IF PCIB.ERR&< >0 THEN ERROR PCIB.BASERR  
! ERROR TRAP  
************************** GWBASIC (National Instruments GPIB Interface) *******************************  
2100 MEASUREMENT$ - SPACE$(20)  
2110  
! DRIVER CAN ONLY READ STRINGS, SO RESERVE  
SPACE IN A STRING  
!
2120 CALL IBRD(MODULE.ADDRESS%, MEASUREMENT$)  
2130 IF IBSTA% &< 0 GOTO 5000  
! TRAP ERROR WITH ERROR HANDLER AT LINE 5000  
2140 MEASURED.VALUE=VAL(MEASUREMENT$)” ! CONVERT THE STRING TO A NUMBER  
Receiving Module Data with BASIC  
20 Introduction To Programming  
Download from Www.Somanuals.com. All Manuals Search And Download.  
***************** Microsoft C (Agilent 82335A/82990A/61062B GPIB Command Library) *********************  
/* Assumes that you have an error handler routine, called ’error_handler’ that accepts a float. The error handler is then  
passed the float that is returned from each call to the library. */  
#include &<stdio.h>  
#include k<chpib.h>  
#include &<cfunc.h>  
#define MODULE_ADDRESS 70501L  
char *cmd;  
float measurement;  
error = ioenter(MODULE_ADDRESS, &&measurement);  
error_handler(error);  
**************************** Microsoft C (National Instruments GPIB Interface) ****************************  
/* Assumes that you have an error handler routine, called ’error_handler’. The error handler is then passed the float that is  
returned from each call to the library. */  
#include &<stdio.h>  
#include &<stdlib.h>  
#include &<decl.h>  
#define ERR (1&<&<15)  
/* Error is detected as bit 15 of ibsta */  
#define STRING_LENGTH 20  
/* Length of string to hold measurement */  
int module_address;  
/* Device is configured in the GPIB.COM handler. Use  
ibfind( ) to assign a value to module_address. */  
char measurement[STRING_LENGTH];  
float measured_value;  
/* Holds float conversion of measurement */  
ibwrt(module_address, measurement, STRING_LENGTH);  
if (ibsta & ERR)  
error_handler();  
measured_value = atof(measurement);  
/* Converts measurement string to float */  
Receiving Module Data with C  
Introduction To Programming 21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3
Language Dictionary  
Introduction  
This section gives the syntax and parameters for all the IEEE 488.2 SCPI commands and the Common commands used by  
the Agilent Series 66l0xA power modules. It is assumed that you are familiar with the material in "Chapter 2 - Introduction  
to Programming". That chapter explains the terms, symbols, and syntactical structures used here and gives an introduction  
to programming.  
The programming commands function the same way in all Agilent Series 66l0xA power modules. Since SCPI syntax  
remains the same for all programming languages, the examples are generic.  
Syntax definitions use the long form, but only short form headers (or "keywords") appear in the examples. If you have any  
concern that the meaning of a header in your program listing will not be obvious at some later time, then use the long form  
to help make your program self-documenting.  
Parameters  
Most commands require a parameter and all queries will return a parameter. The range for a parameter may vary according  
to the model of power module. Parameters for all current models are listed in Table 3-3, at the end of this chapter.  
Related Commands  
Where appropriate, related commands or queries are included. These are listed either because they are directly related by  
function or because reading about them will clarify or enhance your understanding of the original command or query.  
Order of Presentation  
The dictionary is organized as follows:  
IEEE 488.2 common commands, in alphabetical order.  
Subsystem commands.  
COMMON Commands  
Common commands begin with an * and consist of three letters (command) or three letters and a ? (query). Common  
commands are defined by the IEEE 488.2 standard to perform some common interface functions. The Agilent Series  
6610xA power modules respond to the 13 required common commands that control status reporting, synchronization, and  
internal operations. The power modules also respond to five optional common commands controlling triggers, power-on  
conditions, and stored operating parameters.  
Subsystem Commands  
Subsystem commands are specific to power module functions. They can be a single command or a group of commands.  
The groups are comprised of commands that extend one or more levels below the root. The description of subsystem  
commands follows the description of the common commands.  
Language Dictionary 23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Description Of Common Commands  
Figure 3-1 shows the common commands and queries. These commands are listed alphabetically in the dictionary. If a  
command has a corresponding query that simply returns the data or status specified by the command, then both command  
and query are included under the explanation for the command. If a query does not have a corresponding command or is  
functionally different from the command, then the query is listed separately. The description for each common command or  
query specifies any status registers affected. In order to make use of this information, you must refer to "Chapter 4 - Status  
Reporting", which explains how to read specific register bits and use the information that they return.  
Figure 3-1. Common Commands Syntax Diagram  
*CLS  
Meaning and Type  
Clear Status Device Status  
Description  
This command causes the following actions (see "Chapter 4 - Status Reporting" for descriptions of all registers):  
Clears the following registers:  
Standard Event Status  
Operation Status Event  
Questionable Status Event  
Status Byte  
Clears the Error Queue  
24 Language Dictionary  
Download from Www.Somanuals.com. All Manuals Search And Download.  
If *CLS immediately follows a program message terminator (<NL>), then the output queue and the MAV bit  
are also cleared.  
Command Syntax  
Parameters  
*CLS  
(None)  
(None)  
Query Syntax  
*ESE  
Meaning and Type  
Event Status Enable Device Status  
Description  
This command programs the Standard Event Status Enable register bits. The programming determines which events of the  
Standard Event Status Event register (see *ESR?) are allowed to set the ESB (Event Summary Bit) of the Status Byte  
register. A "1" in the bit position enables the corresponding event. All of the enabled events of the Standard Event Status  
Event register are logically ORed to cause the Event Summary Bit (ESB) of the Status Byte register to be set. See "Chapter  
4 - Status Reporting" for descriptions of all three registers.  
Bit Configuration of Standard Event Status Enable Register  
Bit Position  
Bit Name  
Bit Weight  
7
6
0
64  
5
4
EXE  
16  
3
DDE  
8
2
QYE  
4
1
0
2
0
OPC  
1
PON  
128  
CME  
32  
CME = Command error; DDE = Device-dependent error; EXE = Execution error; OPC = Operation  
complete; PON Power-on; QYE = Query error  
Command Syntax  
Parameters  
*ESE <NRf>  
0 to 255  
Power On Value  
Suffix  
(See *PSC)  
(None)  
Example  
*ESE 129  
*ESE?  
<NR1>  
Query Syntax  
Returned Parameters  
Related Commands  
(Register value)  
*ESR? *PSC *STB?  
If PSC is programmed to 0, then the *ESE command causes a write cycle to nonvolatile memory. The  
nonvolatile memory has a finite maximum number of write cycles (see in the power module User’s  
Guide). Programs that repeatedly cause write cycles to nonvolatile memory can eventually  
exceed the maximum number of write cycles and may cause the memory to fail.  
*ESR?  
Meaning and Type  
Event Status Register  
Device Status  
Description  
This query reads the Standard Event Status Event register. Reading the register clears it. The bit configuration of this  
register is the same as the Standard Event Status Enable register (*ESE). See "Chapter 4 - Status Reporting" for a detailed  
explanation of this register.  
Query Syntax  
Parameters  
Returned Parameters  
Related Commands  
*ESR?  
(None)  
<NR1>  
(Register binary value)  
*CLS *ESE *ESE? *OPC  
Language Dictionary 25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
*IDN?  
Identification Query  
Meaning and Type  
Identification System Interface  
Description  
This query requests the power module to identify itself. It returns a string composed of four fields separated by commas.  
Query Syntax  
Returned Parameters  
*IDN?  
<AARD>  
Field  
Information  
Agilent Technologies Manufacturer  
xxxxxA  
5-digit model number followed by a letter  
nnnnA-nnnnn  
<R>.xx.xx  
10-character serial number or 0  
Revision levels of firmware  
Example  
Agilent Technologies,66101A,0,A.00.01  
Related Commands  
(None)  
*OPC  
Meaning and Type  
Operation Complete Device Status  
Description  
This command causes the interface to set the OPC bit (bit 0) of the Standard Event Status register when the power module  
has completed all pending operations. (see *ESE for the bit configuration of the Standard Event Status register.) Pending  
operations are complete when:  
All commands sent before *OPC have been executed. This includes overlapped commands. Most commands  
are sequential and are completed before the next command is executed. Overlapped commands are executed in  
parallel with other commands. Commands that affect output voltage, current or state, relays, and trigger actions are  
overlapped with subsequent commands sent to the power module. The *OPC command provides notification that  
all overlapped commands have been completed.  
Any change in the output level caused by previous commands has been completed (completion of settling  
time, relay bounce, etc.)  
All triggered actions are completed.  
*OPC does not prevent processing of subsequent commands but Bit 0 will not be set until all pending operations are  
completed.  
Command Syntax  
Parameters  
*OPC  
(None)  
Related Commands  
*OPC? *WAI  
*OPC?  
Meaning and Type  
Operation Complete Device Status  
Description  
This query causes the interface to place an ASCII "1" in the Output Queue when all pending operations are completed.  
Pending operations are as defined for the *OPC command. Unlike *OPC, *OPC? prevents processing of all subsequent  
commands. *OPC? is intended to be used at the end of a command line so that the application program can then monitor  
the bus for data until it receives the "1" from the power module Output Queue.  
26 Language Dictionary  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Do not follow *OPC? with *TRG or GPIB bus triggers. Such triggers sent after *OPC? will be  
prevented from executing and will prevent the power module from accepting further commands. If  
this occurs, the only programmable way to restore operation is by sending the power module a GPIB  
DCL (Device Clear) command.  
Query Syntax  
Returned Parameters  
*OPC?  
<NR1>  
ASCII 1 is placed in the Output Queue when the  
power module has completed operations.  
Related Commands  
*OPC *TRIG *WAI  
*OPT?  
Identification Query  
Meaning and Type  
Identification System Interface  
Description  
This query requests the power module to identify any options that are installed. Options are identified by number, as shown  
below. A 0 indicates no options are installed.  
Query Syntax  
Returned Parameters  
Related Commands  
*OPT?  
<AARD>  
(None)  
*PSC  
Meaning and Type  
Power-on Status Clear Device Initialization  
Description  
This command controls the automatic clearing at power on the following registers (see "Chapter 4 - Status Reporting" for  
register details):  
Service Request Enable.  
Standard Event Status Enable.  
If the command parameter = 1 (or any non-zero value), then the above registers are cleared at power on. If the command  
parameter = 0, then the above registers are not cleared but are programmed to their last state prior to power turn on. This is  
the most common application for *PSC and enables the power module to generate an SRQ (Service Request interrupt) at  
power on.  
Command Syntax  
Parameters  
*PSC <bool>  
0 | 1 | OFF | ON  
*PSC 0 *PSC 1  
*PSC?  
Example  
Query Syntax  
Returned Parameters  
Related Commands  
<NR1> 0 | 1  
*ESE *SRE  
*PSC causes a write cycle to nonvolatile memory. If *PSC is programmed to 0, then the *ESE and  
*SRE commands also cause a write cycle to nonvolatile memory. The nonvolatile memory has a finite  
number of write cycles (see Table 1-2 in the power module User’s Guide). Programs that  
repeatedly write to nonvolatile memory can eventually exceed the maximum number of write cycles  
and may cause the memory to fail.  
Language Dictionary 27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
*RCL  
Meaning and Type  
Recall Device State  
Recalling a previously stored state may place hazardous voltage at the power module output.  
Description  
This command restores the power module to a state that was previously stored in memory with a *SAV command to the  
specified location. The following states are recalled:  
CAL:AUTO  
LIST:COUN  
LIST:STEP  
OUTP[:STAT]  
OUTP:DFI[:STAT]  
OUTP:DFI:LINK  
OUTP:PROT:DEL  
OUTP:REL[:STAT]  
OUTP:REL:POL  
OUTP:TTLT[:STAT]  
OUTP:TTLT:LINK  
OUTP:TTLT:SOUR  
TRIG:DEL  
TRIG:LINK  
TRIG:SOUR  
VOLT[:LEV][IMM]  
VOLT:MODE  
VOLT:PROT[:LEV]  
CURR[:LEV][:IMM]  
CURR:MODE  
CURR:PROT:STAT  
DISP:STAT  
INIT:CONT  
Sending *RCL also does the following:  
Forces an ABORt command before resetting any parameters (this cancels any uncompleted trigger actions).  
Disables the calibration function by setting CAL:STATe to OFF.  
The device state stored in location 0 is automatically recalled at power turn-on when the power module configuration switch  
is set for this mode of operation (see the power module User’s Guide).  
Note  
Whenever the power module is powered up, the state stored in location 0 is written to the 5 volatile  
locations (5 through 9).  
Command Syntax  
Parameters  
*RCL <NRf>  
0 through 9  
*RCL 3  
Example  
Query Syntax  
Related Commands  
(None)  
*PSC *RST *SAV  
*RST  
Meaning and Type  
Reset Device State  
Description  
This command resets the power module to a factory-defined state as defined below. *RST also forces an ABORt command.  
COMMAND  
CAL:AUTO  
CAL:STAT  
CURR[:LEV][:IMM]  
CURR:PROT:STAT  
CURR:MODE  
DISP[:WIND]:STAT  
INIT:CONT  
STATE  
OFF  
COMMAND  
OUTP[:STAT]  
OUTP:DFI  
OUTP:DFI:SOUR  
OUTP:DFI:LINK  
OUTP:PROT:DEL  
OUTP:REL[:STAT]  
OUTP:REL:POL  
OUTP:TTLT[STAT]  
OUTP:TTLT:SOUR  
STATE  
OFF  
OFF  
COMMAND  
OUTP.TTLT:LINK  
TRIG:DEL  
TRIG:LINK  
TRIG:SOUR  
VOLT[:LEV][:IMM]  
VOLT:MODE  
VOLT:PROT:LEV  
1Model-dependent value. See  
Table 3-2.  
STATE  
OFF  
0
OFF  
BUS  
0
OFF  
1
LINK  
OFF  
FIX  
ON  
“SUM3”  
1
OFF  
NORM  
OFF  
FIX  
MAX  
OFF  
A UTO  
LIST:STEP  
LIST:COUN  
1
BUS  
28 Language Dictionary  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Command Syntax  
Parameters  
*RST  
(None)  
Query Syntax  
(None)  
Related Commands  
*PSC *SAV  
*SAV  
Meaning and Type  
Save Device State  
Description  
This command stores the present state of the power module to a specified location in memory. Up to 10 states can be  
stored. Storage locations 0 through 4 are in nonvolatile memory and locations 5 through 9 are in volatile memory. If a  
particular state is desired at power on, it should be stored in location 0. It then will be recalled at power on if the power  
module configuration switch is set for this mode of operation (see the power module User’s Guide).  
The following power module states are stored by *SAV:  
CAL:AUTO  
LIST:COUN  
LIST:STEP  
OUTP[:STAT]  
OUTP:DFI[:STAT]  
OUTP:DFI:LINK  
OUTP:PROT:DEL  
OUTP:REL[:STAT]  
OUTP:REL:POL  
OUTP:TTLT[:STAT]  
OUTP:TTLT:LINK  
OUTP:TTLT:SOUR  
TRIG:DEL  
TRIG:LINK  
TRIG:SOUR  
VOLT[:LEV][IMM]  
VOLT:MODE  
VOLT:PROT[:LEV]  
CURR[:LEV][:IMM]  
CURR:MODE  
CURR:PROT:STAT  
DISP:STAT  
INIT:CONT  
Command Syntax  
Parameters  
*SAV  
0 to 9  
Query Syntax  
(None)  
Related Commands  
PSC *RCL *RST  
*SRE  
Meaning and Type  
Service Request Enable Device Interface  
Description  
This command sets the condition of the Service Request Enable register. This register determines which bits from the  
Status Byte register (see *STB for its bit configuration) are allowed to set the Master Status Summary (MSS) bit and the  
Request for Service (RQS) summary bit. A 1 in any Service Request Enable register bit position enables the corresponding  
Status Byte register bit and all such enabled bits then are logically ORed to cause Bit 6 of the Status Byte register to be set.  
See "Chapter 4 - Status Reporting" for more details concerning this process.  
When the controller conducts a serial poll in response to SRQ, the RQS bit is cleared, but the MSS bit is not. When *SRE  
is cleared (by programming it with 0), the power module cannot generate an SRQ to the controller.  
Command Syntax  
Parameters  
*SRE <NRf>  
0 to 255  
Default Value  
(See *PSC)  
*SRE 20  
Example  
Query Syntax  
*SRE?  
Returned Parameters  
Related Commands  
<NR1>  
*ESE *ESR *PSC  
(Register binary value)  
Language Dictionary 29  
Download from Www.Somanuals.com. All Manuals Search And Download.  
If *PSC is programmed to 0, then the *SRE command causes a write cycle to nonvolatile memory.  
The nonvolatile memory has a finite number of write cycles (see Table 1-2 in the power module  
User’s Guide). Programs that repeatedly write to nonvolatile memory can eventually exceed the  
maximum number of write cycles and may cause the memory to fail.  
*STB?  
Meaning and Type  
Status Byte Device Status  
Description  
This query reads the Status Byte register, which contains the status summary bits and the Output Queue MAV bit. Reading  
the Status Byte register does not clear it. The input summary bits are cleared when the appropriate event registers are read  
(see “Chapter 4 - Status Reporting”) for more information). The MAV bit is cleared at power on or by *CLS.  
A serial poll also returns the value of the Status Byte register, except that bit 6 returns Request for Service (RQS) instead of  
Master Status Summary (MSS). A serial poll clears RQS, but not MSS. When MSS is set, it indicates that the power  
module has one or more reasons for requesting service.  
Bit Configuration of Status Byte Register  
Bit Position  
Condition  
7
6
5
ESB  
4
3
2
1
0
2
2
2
OPER  
MSS  
(RQS)  
64  
MAV  
QUES  
Bit Weight  
128  
32  
16  
8
4
2
1
ESB = Event status byte summary; M = Message available  
MSS = Master status summary; OPER = Operation status summary;  
QUES = Questionable status summary; RQS = Request for service  
1Also represents RQS. 2These bits are always zero.  
Query Syntax  
Returned Parameters  
Related Commands  
*STB?  
<NR1>  
(None)  
(Register binary value)  
*TRG  
Meaning and Type  
Trigger Device Trigger  
Description  
This command generates a trigger to any subsystem that has BUS selected as its source (for example, TRIG:SOUR BUS,  
OUTP:TTLT:SOUR BUS). The command has the same affect as the Group Execute Trigger (<GET>) command.  
Command Syntax  
Parameters  
*TRG  
(None)  
Query Syntax  
(None)  
Related Commands  
ABOR CURR:TRIG INIT TRIG[:IMM] VOLT:TIUG  
30 Language Dictionary  
Download from Www.Somanuals.com. All Manuals Search And Download.  
*TST?  
Meaning and Type  
Test Device Test  
Description  
This query causes the power module to do a self-test and report any errors (see "Selftest Error Messages" in Chapter 3 of the  
power module User’s Guide).  
Query Syntax  
Returned Parameters  
*TST?  
<NR1>  
0
Indicates power module passed self-test.  
Indicates an error code.  
Nonzero  
(None)  
Related Commands  
*WAI  
Meaning and Type  
Wait to Continue Device Status  
Description  
This command instructs the power module not to process any further commands until all pending operations are completed.  
"Pending operations" are as defined under the *OPC command. *WAI can be aborted only by sending the power module a  
GPIB DCL (Device Clear) command.  
Command Syntax  
Parameters  
Query Syntax  
Related Commands  
*WAI  
(None)  
(None)  
*OPC  
Description of Subsystem Commands  
Figure 3-2 is a tree diagram of the subsystem commands. Commands followed by a question mark (?) take only the query  
form. Except as noted in the syntax descriptions, all other commands take both the command and query form. The  
commands are listed in alphabetical order and the commands within each subsystem are grouped alphabetically under the  
subsystem.  
ABOR  
This command cancels any trigger actions presently in process. Pending trigger levels are reset equal to their corresponding  
immediate values. ABOR also cancels any programmed lists that may be in process.  
ABOR also resets the WTG bit in the Operation Condition Status register (see "Chapter 4 - Status Reporting"). If  
INIT:CONT ON has been programmed, the trigger subsystem initiates itself immediately afterABORt, thereby setting  
WTG. ABOR is executed at power turn on and upon execution of *RCL, RST, or any implied abort command (see List  
Subsystem).  
Command Syntax ABORt  
Parameters (None)  
Examples ABOR  
Query Syntax (None)  
Related Commands INIT *RST *TRG TRIG  
Language Dictionary 31  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Figure 3-2. Subsystem Tree Diagram  
Calibration Subsystem  
The commands in this subsystem allow you to do the following:  
Control automatic calibration of the measurement subsystem.  
Enable and disable the calibration mode.  
Change the calibration password.  
Calibrate the overvoltage protection (OVP) circuit.  
Calibrate the current and voltage output levels, and store new calibration constants in nonvolatile memory.  
CAL:AUTO  
This command controls the autocalibration function and is used to substantially improve the accuracy of the  
MEAS:CURR? and MEAS:VOLT? data readback queries. It does this by compensating for temperature drift in the  
readback circuitry.  
32 Language Dictionary  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Whenever CAL:AUTO ONCE is sent, the power module performs an immediate readback temperature compensation.  
CAL:AUTO ONCE is a sequential command that takes several seconds to complete. When CAL:AUTO ON is sent, the  
power module automatically performs a readback temperature compensation before executing every MEAS command. Use  
of this command extends the execution time of every MEAS query.  
Command Syntax  
Parameters  
CALibrate:AUTO <bool> | ONCE  
0 | OFF | 1 | ON | ONCE  
OFF  
*RST Value  
Examples  
Query Syntax  
Returned Parameters  
Related Commands  
CAL:AUTO 1 CAL:AUTO ONCE  
CALibrate:AUTO?  
0 | 1  
MEAS:CURR? MEAS:VOLT?  
CAL:CURR  
This command can only be used in the calibration mode. It enters a current value that you obtain by reading an external  
meter. You must first select a calibration level (CAL:CURR:LEV) for the value being entered. Two successive values  
(one for each end of the calibration range) must be selected and entered. The power module then computes new current  
calibration constants. These constants are not stored in nonvolatile memory until saved with the CAL:SAVE command.  
Command Syntax  
Parameters  
CALibrate:CURRent[:DATA] <NRf>  
(See Table 3-2)  
Default Suffix  
Examples  
Query Syntax  
Related Commands  
A
CAL:CURR 3222.3 MA CAL:CURR:DATA 5.000  
(None)  
CAL:SAVE CAL:STAT  
CAL:CURR:LEV  
This command can only be used in the calibration mode. It sets the power module to a calibration point that is then entered  
with CAL:CURR[:DATA]. During calibration, two points must be entered and the low-end point (MIN) must be selected  
and entered first.  
Command Syntax  
Parameters  
CALibrate:CURRent:LEVel <CRD>  
MINimum |MAXimum  
Examples  
Query Syntax  
Related Commands  
CAL:CURR:LEV MIN CAL:CURR:LEV MAX  
(None)  
CAL:CURR[:DATA] CAL:STAT  
CAL:PASS  
This command can only be used in the calibration mode. It allows you to change the calibration password. Unless it is  
changed subsequently to shipment, the password is the model number of the power module. A new password is  
automatically stored in nonvolatile memory and does not have to be stored with the CAL:SAVE command.  
If the password is set to 0, password protection is removed and the ability to enter the calibration mode is unrestricted.  
Command Syntax CALibrate:PASScode <NRf>  
Parameters <NRf>  
Examples CAL:PASS 66102 CAL:PASS 09.1991  
Query Syntax (None)  
Related Commands CAL:STAT  
Language Dictionary 33  
Download from Www.Somanuals.com. All Manuals Search And Download.  
CAL:SAVE  
This command can only be used in the calibration mode. It saves any new calibration constants (after a current or voltage  
calibration procedure has been completed) in nonvolatile memory.  
Command Syntax: CALibrate:SAVE  
Parameters (None)  
Examples CAL:SAVE  
Query Syntax (None)  
Related Commands CAL:CURR CAL:VOLT CAL:STAT  
CAL:STAT  
This command enables and disables the calibration mode. The calibration mode must be enabled before the power module  
will accept any other calibration commands except CAL:AUTO.  
The first parameter specifies the enabled or disabled state. The second parameter is the password. It is required if the  
calibration mode is being enabled and the existing password is not 0. If the second parameter is not entered or is incorrect,  
an error is generated and the calibration mode remains disabled. The query statement returns only the state, not the  
password.  
Whenever the calibration mode is changed from enabled to disabled, any new calibration constants are lost unless they have  
been stored with CAL:SAVE.  
Command Syntax: CALibrate:STATe <bool> [,<NRf>]  
Parameters 0 | OFF | 1 | ON[,<NRf>]  
*RST Value OFF  
Examples CAL:STAT 1,66102 CAL:STAT OFF  
Query Syntax CALibrate:STATe?  
Returned Parameters 0 | 1  
Related Commands CAL:PASS CAL:SAVE  
CAL:VOLT  
This command can only be used in the calibration mode. It enters a voltage value that is obtained from an external meter.  
You must first select a calibration level (CAL:VOLT:LEV) for the value being entered. Two successive values (one for  
each end of the calibration range) must be selected and entered. The power module then computes new voltage calibration  
constants. These constants are not stored in nonvolatile memory until saved with the CAL:SAVE command.  
Command Syntax  
Parameters  
CALibrate:VOLTage[:DATA] <NRf>  
See Table 3-2  
Default Suffix  
Examples  
Query Syntax  
Related Commands  
V
CAL:VOLT 310.0 MV CAL:VOLT 5.000  
(None)  
CAL:SAVE CAL:STAT  
CAL:VOLT:LEV  
This command can only be used in the calibration mode. It sets the power module to a calibration point that is then entered  
with CAL:VOLT[:DATA]. During calibration, two points must be entered and the low-end point (MIN) must be selected  
and entered first.  
Command Syntax  
Parameters  
CALibrate:VOLTage:LEVel <CRD>  
MINimum |MAXimum  
Examples  
Query Syntax  
Related Commands  
CAL:VOLT:LEV MIN CAL:VOLT:LEV MAX  
(None)  
CAL:VOLT[:DATA] CAL:STAT  
34 Language Dictionary  
Download from Www.Somanuals.com. All Manuals Search And Download.  
CAL:VOLT:PROT  
This command can only be used in the calibration mode. It calibrates the power module overvoltage protection (OV)  
circuit. The power module output must be enabled and operating in the constant voltage (CV) mode. The power module  
automatically performs the calibration and stores the new OV constant in nonvolatile memory. CAL:VOLT:PROT is a  
sequential command that takes several seconds to complete.  
Command Syntax:  
Parameters  
CALibrate:VOLTage:PROTection  
(None)  
Example  
Query Syntax  
Related Commands  
CAL:VOLT:PROT  
(None)  
CAL:STAT  
Current Subsystem  
This subsystem programs the output current of the power module.  
CURR  
This command directly programs the immediate current level of the power module. The immediate level is the current  
applied at the output terminals. This command is always active, even when the current subsystem is in the list mode (see  
CURR:MODE).  
Command Syntax  
Parameters  
[SOURce]:CURRent[:LEVel] [:IMMediate][:AMPLitude] <NRf+>  
See Table 3-2  
Default Suffix  
*RST Value  
Examples  
A
See Table 3-2  
CURR 500 MA CURR:LEV .5  
Query Syntax  
[SOURce]:CURRent[:LEVel] [:IMMediate][:AMPLitude]?  
[SOURce]:CURRent[:LEVel] [:IMMediate][:AMPLitude]? MAX  
[SOURce]:CURRent[:LEVel] [:IMMediate][:AMPLitude]? MIN  
<NR3> CURR? returns the present programmed current level.  
CURR? MAX and CURR? MIN return the maximum and minimum  
programmable current levels.  
Returned Parameters  
Related Commands  
*SAV *RCL *RST  
CURR:MODE  
This command enables or disables list subsystem control over the power module output current. When programmed with  
FIX, this command prevents the output current from being controlled by the sequencing of points specified by  
LIST:CURR. If the LIST parameter is used, then the output current may be changed by the subsequent execution of a list.  
However, the list mode does not prevent the output current from being set by CURR and *RCL.  
Note  
CURR:MODE:LIST is an implied ABORt command.  
Command Syntax [SOURce]:CURRent:MODE <CRD>  
Parameters FIXed | LIST  
*RST Value FIX  
Examples CURR:MODE LIST CURR:MODE FIX  
Query Syntax [SOURce]:CURRent:MODE?  
Returned Parameters FIX | LIST  
Related Commands CURR:LIST *RCL  
Language Dictionary 35  
Download from Www.Somanuals.com. All Manuals Search And Download.  
CURR:PROT:STAT  
This command enables or disables the power module overcurrent (OC) protection function. If the overcurrent protection  
function is enabled and the power module goes into constant current (CC) mode, then the output is disabled and the  
Questionable Condition status register OC bit is set (see "Chapter 4 - Status Reporting"). An overcurrent condition can be  
cleared with the OUTP:PROT:CLE command after the cause of the condition is removed.  
Command Syntax  
Parameters  
[SOURce]:CURRent:PROTection:STATe <bool>  
0 | 1 | OFF | ON  
*RST Value  
OFF  
Examples  
Query Syntax  
Returned Parameters  
Related Commands  
CURR:PROT:STAT 0 CURR:PROT:STAT OFF  
[SOURce]:CURRent:PROTection:STATe?  
0 | 1  
OUTP:PROT:CLE OUTP:PROT:DEL *RCL *SAV  
CURR:TRIG  
This command programs the pending triggered current level of the power module. The pending triggered current level is a  
stored value that is transferred to the output terminals when a trigger occurs. A pending triggered level is unaffected by  
subsequent CURR commands and remains in effect until the trigger subsystem receives a trigger or an ABORt command is  
given. If there is no pending triggered level, then the query form returns the IMMediate current level. In order for  
CURR:TRIG to be executed, the trigger subsystem must be initiated (see INITiate).  
Command Syntax  
Parameters  
[SOURce]:CURRent[:LEVel]:TRIGgered [:AMPLitude] <NRf+>  
See Table 3-2  
Default Suffix  
*RST Value  
Examples  
A
See Table 3-2  
CURR:TRIG 1200 MA CURR:LEV:TRIG 1.2  
Query Syntax  
SOURce]:CURRent[LEVel]:TRIGgered [:AMPLitude]?  
[SOURce]:CURRent[LEVel]:TRIGgered [:AMPLitude]? MAXimum  
[SOURce]:CURRent[:LEVel]:TRIGgered [:AMPLitude]? MIN  
<NR3> CURR:TRIG? returns the presently programmed triggered level.  
If no triggered level is programmed, the CURR level is returned.  
CURR:TRIG? MAX and CURR:TRIG? MIN return the maximum and  
minimum programmable triggered current levels.  
ABOR CURR[:IMM] CURR:MODE INIT *RST  
Returned Parameters  
Related Commands  
DISPlay Command  
This command turns the power module optional front panel voltage and current displays on and off. It does not affect the  
annunciators.  
Command Syntax DISPlay[:WINDow]:STAT <bool>  
Parameters 0 | 1 | OFF | ON  
*RST Value ON  
Examples DISP:STAT 1 DISP:STAT OFF  
Query Syntax DISPlay[:WINDow]:STAT?  
Returned Parameters 0 |1  
Related Commands *SAV *RCL  
INITiate Command  
This command enables the trigger subsystem. When a trigger is enabled, an event on the selected trigger source causes the  
specified triggering action to occur. If a trigger circuit is not enabled, all trigger commands are ignored. If INIT:CONT is  
OFF, then INIT enables the trigger subsystem only for a single trigger action. The subsystem must be enabled prior to each  
subsequent trigger action. If INIT:CONT is ON, then the trigger system is continuously enabled and INIT is redundant.  
36 Language Dictionary  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Command Syntax INITiate[:IMMediate]  
INITiate:CONTinuous <bool>  
Parameters For INIT[:IMM]  
For INIT:CONT  
(None)  
0 | 1 | OFF | ON  
*RST Value OFF  
Examples INIT INIT:CONT 1 INIT:CONT ON  
Query Syntax For INIT[:IMM]  
For INIT:CONT  
(None)  
INITiate:CONTinuous?  
Returned Parameters 0 | 1  
Related Commands ABOR CURR:TRIG TRIG *TRG VOLT:TRIG  
List Subsystem  
This subsystem controls the generation of parameter lists that sequence the power module output through values of voltage  
and current. Two subsystem commands specify lists of output voltages (LIST:VOLT), and currents (LIST:CURR). A  
count command (LIST:COUN) determines how many times the power module sequences through a list before that list is  
completed. A dwell command (LIST:DWEL) specifies the time interval that each value (point) of a list is to remain in  
effect. A step command (LIST:STEP) determines if a trigger causes a list to advance only to its next point or to sequence  
through all of its points.  
Each list can have from 1 to 20 points. Normally, voltage, current, and dwell lists must have the same number of points, or  
an error is generated when the first list point is triggered. The exception is a list consisting of only one point. Such a list is  
treated as if it had the same number of points as the other lists, with all the points having the same value as the one specified  
point.  
Note  
All list subsystem commands (as well as CURR:MODE LIST and VOLT:MODE LIST) are implied  
ABORt commands.  
LIST:COUN  
This command sets the number of times that the list is executed before it is completed. The command accepts parameters in  
the range 1 through 9.9E37, but any number greater than 65534 is interpreted as INFinity. Use INF if you wish to execute a  
list indefinitely.  
Command Syntax  
Parameters  
[SOURce]:LIST:COUNt <NRf+>  
1 to 9.9E37 | INFinity  
*RST Value  
1
Examples  
Query Syntax  
Returned Parameters  
Related Commands  
LIST:COUN 3 LIST:COUN INF  
[SOURce]:LIST:COUNt?  
<NR3>  
CURR:MODE LIST:CURR LIST:DWEL LIST:STEP  
LIST:VOLT VOLT:MODE  
LIST:CURR  
This command specifies the output current points in a list. The current points are given in the command parameters, which  
are separated by commas. Up to 20 points may be entered and the output current values specified by the points will be  
generated in the same order as they were entered.  
Command Syntax  
Parameters  
[SOURce]:LIST:CURRent <NRf+> {,<NRf+>}  
See Table 3-2  
Default Suffix  
Examples  
Query Syntax  
Related Commands  
A
LIST:CURR 2.5,3.0,3.5 LIST:CURR MAX,2.5,MIN  
(None)  
CURR:MODE LIST:CURR:POIN? LIST:DWEL  
Language Dictionary 37  
Download from Www.Somanuals.com. All Manuals Search And Download.  
LIST:CURR:POIN?  
This query returns the number of points specified in LIST:CURR. Note that it returns only the total number of points, not  
the point values.  
Query Syntax  
Returned Parameters  
Example  
[SOURce]:LIST:CURRent:POINts?  
<NR1>  
LIST:CURR:POIN?  
Related Commands  
CURR:MODE LIST:CURR LIST:DWEL  
LIST:DWEL  
This command sets the dwell points for the output current list and output voltage list. Each dwell point specifies the time, in  
seconds, that the output of the power module is to remain at the level specified by the corresponding point in the current or  
voltage list. At the end of the dwell time, the output of the power module depends upon the following conditions:  
If LIST:STEP AUTO has been programmed, the output automatically changes to the next point in the list.  
If LIST:STEP ONCE has been programmed, the output remains at the present level until a trigger sequences  
the next point in the list.  
Command Syntax  
Parameters  
[SOURce]:LIST:DWEL1 <NRf+> {,<NRf+>}  
0.01 to 65 |MINimum | MAXimum  
Default Suffix  
Examples  
S
LIST:DWEL .5,.5,1.5  
Query Syntax  
Related Commands  
(None)  
CURR:MODE LIST:COUN LIST:CURR LIST:STEP  
LIST:VOLT VOLT:MODE  
LIST:DWEL:POIN?  
This query returns the number of points specified in LIST:DWEL. Note that it returns only the total number of points, not  
the point values.  
Query Syntax  
Returned Parameters  
Example  
[SOURce]:LIST:DWFL1:POINts?  
<NR1>  
LIST:DWEL:POIN?  
Related Commands  
LIST:CURR LIST:DWEL LIST:VOLT  
LIST:STEP  
This command specifies how list sequencing occurs in response to triggers. If LIST:STEP AUTO is sent, then a single  
trigger causes the list (voltage, current, or dwell) to sequence through all its points. The time that a list remains at each  
point is as specified in the dwell list. As soon as the dwell interval expires, the list moves to the next point.  
If LIST:STEP ONCE is sent, then a single trigger advances a list only one point. After the specified dwell interval, the list  
remains at that point until the next trigger occurs.  
In either mode, triggers that occur during a dwell interval are ignored.  
Command Syntax  
Parameters  
[SOURce]:LIST:STEP <CRD>  
AUTO | ONCE  
*RST Value  
AUTO  
Examples  
LIST:STEP ONCE  
Query Syntax  
Returned Parameters  
Related Commands  
[SOURce]:LIST:STEP?  
AUTO | ONCE  
CURR:MODE LIST:COUN LIST:CURR LIST:DWEL  
LIST:VOLT VOLT:MODE  
38 Language Dictionary  
Download from Www.Somanuals.com. All Manuals Search And Download.  
LIST:VOLT  
This command specifies the output voltage points in a list. The voltage points are given in the command parameters, which  
are separated by commas. Up to 20 points may be entered and the output voltage values specified by the points will be  
generated in the same order as they were entered.  
Command Syntax  
Parameters  
[SOURce]:LIST:VOLTage <NRf+> {,<NRf+>}  
See Table 3-2  
Default Suffix  
Examples  
Query Syntax  
Related Commands  
V
LIST:VOLT 2.0,2.5,3.0 LIST:VOLT MAX,2.5,MIN  
(None)  
VOLT:MODE LIST:VOLT:POIN? LIST:DWEL  
LIST:VOLT:POIN?  
This query returns the number of points specified in LIST:VOLT. Note that it returns only the total number of points, not  
the point values.  
Query Syntax  
Returned Parameters  
Example  
[SOURce]:LIST:VOLTage:POINts?  
<NR1>  
LIST:VOLT:POIN?  
Related Commands  
VOLT:MODE LIST:VOLT LIST:DWEL  
MEASure Query  
This query returns the current measured at the power module output terminals or the voltage measured at the sense  
terminals. The query format allows two optional parameters for specifying the expected value and desired measurement  
accuracy. The power module accepts the optional parameters but ignores them.  
Query Syntax  
MEASure:CURRent[:DC]? [<NRf>[,<NRf>]]  
MEASure:VOLTage[:DC]? [<NRf>[,<NRf>]]  
(None)  
A for MEAS:CURR? V for MEAS:VOLT?  
MEAS:CURR? MEAS:VOLT?  
MEASURE:VOLTAGE:DC? MV  
<NR3>  
Parameters  
Default Suffix  
Examples  
Returned Parameters  
Related Commands  
CAL:AUTO  
Output Subsystem  
This subsystem controls the power module voltage and current outputs and the optional output relay.  
OUTP  
This command enables or disables the power module output. The state of a disabled output is a condition of zero output  
voltage and a model-dependent minimum source current.  
If the power module is configured to use the relay option, the command opens the relay contacts when the output is disabled  
and closes them when it is enabled. Transitions between the output ON and OFF states are sequenced so that the relay is  
switched while the power mesh is disabled. Use of the second (NORelay) parameter prevents the command from having  
any effect on the relay; it remains in its existing state whenOUTPut is executed. The query form returns the output state,  
excluding that of the relay (see OUTP:REL?).  
Language Dictionary 39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Command Syntax  
Parameters  
OUTPut[:STATe] <bool>[,NORelay]  
0 | OFF[,NORelay] | 1 | ON[,NORelay]  
*RST Value  
0
Examples  
Query Syntax  
Returned Parameters  
Related Commands  
OUTP 1 OUTP:STAT ON,NORELAY  
OUTPut[:STATe]?  
0 | 1  
*RCL *SAV  
OUTP:DFI  
This command enables or disables the discrete fault indicator (DFI) signal to the power module backplane.  
Command Syntax  
Parameters  
OUTPut:DFI[:STATe] <bool>  
0 |1 | OFF | ON  
*RST Value  
OFF  
Examples  
Query Syntax  
Returned Parameters  
Related Commands  
OUTP:DFI:1 OUTP:DFI OFF  
OUTPut:DFI[:STATe]?  
0 | 1  
OUTP:DFI:LINK OUTP:DFI:SOUR  
OUTP:DFI:LINK  
This command specifies which events within the power module are linked to DFI source events.  
Command Syntax  
Parameters  
OUTPut:DFI:LINK <CRD>  
See Table 3-2  
*RST Value  
SUM3  
Examples  
Query Syntax  
Returned Parameters  
Related Commands  
OUTP:DFI:LINK "CC" OUTP:DFI:LINK OFF”  
OUTPut:DFI:LINK?  
See Table 3-2  
OUTP:DFI:SOUR OUTP:DFI[:STAT]  
OUTP:DFI:SOUR  
This command selects the source for DFI events. The only available source is LINK.  
Command Syntax  
Parameters  
OUTP:DFI:SOUR <CRD>  
LINK  
*RST Value  
LINK  
Examples  
Query Syntax  
Returned Parameters  
Related Commands  
OUTP:DFI:SOUR LINK  
OUTPut:DFI:SOUR?  
LINK  
OUTP:DFI:LINK OUTP:DFI[:STAT]  
OUTP:PROT  
There are two output protection commands that do the following:  
OUTP:PROT:CLE Clears any overvoltage (OV), overcurrent (OC), overtemperature (OT), or remote inhibit (RI)  
protection features. After this command, the output is restored to the state it was in before the  
protection feature occurred.  
40 Language Dictionary  
Download from Www.Somanuals.com. All Manuals Search And Download.  
OUTP:PROT:DEL  
Sets the delay time between the programming of an output change that produces a CV, CC, or  
UNREG condition and the recording of that condition by the Status Operation Condition register.  
The delay prevents momentary changes in power module status that can occur during  
reprogramming from being registered as events by the status subsystem. Since the delay applies  
to CC status, it also delays the OCP (overcurrent protection) feature. The OVP (overvoltage  
protection) feature is not affected by this delay.  
Command Syntax  
Parameters  
OUTPut:PROTection:CLEar  
OUTPut:PROTection:DELay <NRf+>  
OUTP:PROT:CLE, (none)  
OUTP:PROT:DEL 0 to 32.767 | MIN | MAX  
S
Default Suffix  
*RST Value  
Examples  
100 (milliseconds)  
OUTP:PROT:CLE OUTP:PROT:DEL 75E-1  
OUTP:PROTection:CLEar (None)  
OUTPut:PROTection:DELay?  
OUTPut:PROTection:DELay? MINimum  
OUTPut:PROTection:DELay? MAXimum  
<NR3>  
Query Syntax  
Returned Parameters  
Related Commands  
OUTP:PROT:CLE (None)  
OUTP:PROT:DEL *RCL *SAV  
OUTP:REL  
This command is valid only if the power module is configured for the optional relay connector. Programming ON closes  
the relay contacts; programming OFF opens them. The relay is controlled independently of the output state. If the power  
module is supplying power to a load, that power will appear at the relay contacts during switching. If the power module is  
not configured for the relay connector, sending either relay command generates an error.  
Command Syntax OUTPut:RELay[:STATe] <bool>  
Parameters 0 | 1 | OFF | ON  
*RST Value  
0
Examples OUTP:REL 1 OUTP:REL OFF  
Query Syntax OUTPut:RELay?  
Returned Parameters 0 |1  
Related Commands OUTP[:STAT] *RCL *SAV  
OUTP:REL:POL  
This command is valid only if the power module is configured for the optional relay connector. Programming NORMal  
causes the relay output polarity to be the same as the power module output. Programming REVerse causes the relay output  
polarity to be opposite to that of the power module output. If OUTP[:STAT] = ON when either relay command is sent, the  
power module output voltage is set to 0 during the time that the relays are changing polarity. If the power module is not  
configured for the relay connector, sending either relay command generates an error.  
Command Syntax  
Parameters  
OUTPut:RELay:POLarity <CRD>  
NORMal | REVerse  
*RST Value  
NORM  
Examples  
Query Syntax  
Returned Parameters  
Related Commands  
OUTP:REL:POL NORM  
OUTPut:RELay:POLarity?  
NORM | REV  
OUTP[:STAT] *RCL *SAV  
Language Dictionary 41  
Download from Www.Somanuals.com. All Manuals Search And Download.  
OUTP:TTLT  
This command enables or disables the power module Trigger Out signal, which is available at a BNC connector on the rear  
of the mainframe. Trigger Out is the logical OR of all the power module TTLTrig signals (see "Chapter 5 - Synchronizing  
Power Module Output Changes"). It also may be selected as a trigger input (see TRIGger:SOURce).  
Command Syntax OUTPut:TTLTrg[:STATe] <bool>  
Parameters 0 | 1 | OFF | ON  
*RST Value OFF  
Examples OUTP:TTLT 1 OUTP:TTLT OFF  
Query Syntax OUTPut:TTLrg[:STATe]?  
Returned Parameters 0 |1  
Related Commands OUTP:TTLT:LINK OUTP:TTLT:SOUR  
OUTP:TTLT:LINK  
This command specifies which events within the power module are linked to TTLTrg source events when LINK is the  
parameter for the OUTP:TTLT:SOUR command.  
Command Syntax  
Parameters  
OUTPut:TTLrg:LINK <CRD>  
See Table 3-1  
*RST Value  
OFF  
Examples  
Query Syntax  
Returned Parameters  
Related Commands  
OUTP:TTLT:LINK "CC" OUTP:TTLT:LINK "OFF"  
OUTPut:TTLrg:LINK?  
See Table 3-1  
OUTP:TTLT:SOUR OUTP:TTLT[:STAT]  
OUTP:TTLT:SOUR  
This command selects the signal source for the Trig Out signal as follows:  
BUS  
EXT  
*TRG or <GET> (Group Execute Trigger)  
Mainframe backplane Trigger In bus  
HOLD  
LINK  
No trigger source except TRIG:IMM  
Internal power module event as  
specified by TRIG:LINK  
When an event becomes true at the selected TTLTrg source, a pulse is sent to the BNC connector on the rear of the  
mainframe.  
Command Syntax  
Parameters  
OUTPut:TTLrg:SOURce <CRD>  
BUS | EXTernal | LINK | HOLD  
BUS  
*RST Value  
Examples  
OUTP:TTLT:SOUR LINK  
Query Syntax  
Returned Parameters  
Related Commands  
OUTPut:TTLrg:SOURce?  
BUS | EXT | LINK | HOLD  
OUTP:TTLT:LINK OUTP:TTLT[:STAT]  
Status Subsystem  
This subsystem programs the power module status registers. The power module has three groups of status registers;  
Operation, Questionable, and Standard Event. The Standard Event group is programmed with Common commands as  
described in "Chapter 4 - Status Reporting". The Operation and Questionable status groups each consist of the following  
five registers:  
Condition  
Enable Event NTR Filter  
PTR Filter  
42 Language Dictionary  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Status Operation Registers  
The bit configuration of all Status Operation registers is shown in the following table:  
Bit Configuration of Operation Registers  
12  
11  
10  
CC  
9
NU  
8
CV  
256  
7
NU  
128  
6
NU  
64  
5
4
3
NU  
8
2
NU  
4
1
NU  
2
0
Bit Position  
Bit Name  
Bit Weight  
STC NU  
4096 2048 1024 512  
WTG NU  
32 16  
CAL  
1
CAL = Interface is computing new calibration constants; CC = The power module is in constant current mode;  
CV = The power module is in constant voltage mode; NU = (Not used); STC = The list step is complete;  
WTG = Interface is waiting for a trigger.  
Note  
See "Chapter 4 - Status Reporting" for more explanation of these registers  
STAT:OPER?  
This query returns the value of the Operation Event register. The Event register is a read-only register which holds (latches)  
all events that are passed by the Operation NTR and/or PTR filter. Reading the Operation Event register clears it.  
Query Syntax  
Parameters  
STATus:OPERation[:EVENt]?  
(None)  
Returned Parameters  
Examples  
<NR1> (Register Value)  
STAT:OPER:EVEN?  
Related Commands  
*CLS STAT:OPER:NTR STAT:OPER:PTR  
STAT:OPER:COND?  
This command returns the value of the Operation Condition register. That is a read-only register which holds the real-time  
(unlatched) operational status of the power module.  
Query Syntax  
Parameters  
STATus:OPERation:CONDition?  
(None)  
Examples  
Returned Parameters  
Related Commands  
STAT:OPER:COND?  
<NR1> (Register value)  
(None)  
STAT:OPER:ENAB  
This command and its query set and read the value of the Operation Enable register. This register is a mask for enabling  
specific bits from the Operation Event register to set the operation summary bit (OPER) of the Status Byte register. This bit  
(bit 7) is the logical OR of all the Operation Event register bits that are enabled by the Status Operation Enable register.  
Command Syntax STATus:OPERation:ENABle <NRf>  
Parameters 0 to 32727  
Suffix (None)  
Default Value  
0
Examples STAT:OPER:ENAB 1312 STAT:OPER:ENAB 1  
Query Syntax STATus:OPERation:ENABle?  
Returned Parameters <NRI> (Register value)  
Related Commands STAT:OPER:EVEN  
Language Dictionary 43  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STAT:OPER:NTR|PTR Commands  
These commands set or read the value of the Operation NTR (Negative-Transition) and PTR (Positive-Transition) registers.  
These registers serve as polarity filters between the Operation Enable and Operation Event registers to cause the following  
actions:  
When a bit in the Operation NTR register is set to 1, then a 1-to-0 transition of the corresponding bit in the  
Operation Condition register causes that bit in the Operation Event register to be set.  
When a bit of the Operation PTR register is set to 1, then a 0-to-I transition of the corresponding bit in the  
Operation Condition register causes that bit in the Operation Event register to be set.  
If the same bits in both NTR and PTR registers are set to 1, then any transition of that bit at the Operation  
Condition register sets the corresponding bit in the Operation Event register.  
If the same bits in both NTR and PTR registers are set to 0, then no transition of that bit at the Operation  
Condition register can set the corresponding bit in the Operation Event register.  
Note  
Setting a bit in the value of the PTR or NTR filter can of itself generate positive or negative events in the  
corresponding Operation Event register.  
Command Syntax  
STATus:OPERation:NTRansition <Nrf>  
STATus:OPERation:PTRansition <NRf>  
0 to 32727  
Parameters  
Suffix  
(None)  
Default Value  
Examples  
Query Syntax  
Returned Parameters  
Related Commands  
0
STAT: OPER: NTR 32 STAT: OPER: PTR 1312  
STAT:OPER:NTR? STAT:OPER:PTR?  
<NR1> (Register value)  
STAT:OPER:ENAB  
STAT:PRES  
This command sets all defined bits in the Status Subsystem PTR registers and clears all bits in the subsystem NTR and  
Enable registers. STAT:OPER:PTR is set to 1313 and STAT:QUES:PTR is set to 1555.  
Command Syntax  
Parameters  
STATus:PRESet  
(None)  
Examples  
Query Syntax  
Related Commands  
STAT:PRES  
(None)  
(None)  
Status Questionable Registers  
The bit configuration of all Status Questionable registers is as follows:  
Bit Configuration of Questionable Registers  
15-11 10  
9
8
NU  
256  
7
NU  
128  
6
NU  
64  
5
NU  
32  
4
OT  
16  
3
NU  
8
2
NU  
4
1
OC  
2
0
OV  
1
Bit Position  
Condition  
Bit Weight  
NU  
UNR RI  
1024 512  
NU = (Not used); OC = Overcurrent protection circuit has tripped; OT = Overtemperature status condition exists;  
OV = Overvoltage protection circuit has tripped; RI = Remote inhibit is active; UNR = Power supply output is  
unregulated.  
Note  
See "Chapter 4 - Status Reporting" for more explanation of these registers.  
44 Language Dictionary  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STAT:QUES?  
This command returns the value of the Questionable Event register. The Event register is a read-only register which holds  
(latches) all events that are passed by the Questionable NTR and/or PTR filter. Reading the Questionable Event register  
clears it.  
Query Syntax  
Parameters  
STATus:QUEStionable[:EVENt]?  
(None)  
Returned Parameters  
Examples  
<NR1> (Register Value)  
STAT:QUES:EVEN?  
Related Commands  
*CLS STAT:QUES:NTR STAT:QUES:PTR  
STAT:QUES:COND?  
This query returns the value of the Questionable Condition register. That is a read-only register which holds the real-time  
(unlatched) questionable status of the power module.  
Query Syntax  
Example  
STATus:QUEStionable:CONDition?  
STAT: QUES: COND?  
Returned Parameters  
Related Commands  
<NR1>  
(None)  
(Register value)  
STAT:QUES:ENAB  
This command sets or reads the value of the Questionable Enable register. This register is a mask for enabling specific bits  
from the Questionable Event register to set the questionable summary (QUES) bit of the Status Byte register. This bit (bit  
3) is the logical OR of all the Questionable Event register bits that are enabled by the Questionable Status Enable register.  
Command Syntax  
Parameters  
STATus:QUEStionable:ENABle <NRf>  
0 to 32727  
Suffix  
(None)  
Default Value  
Example  
0
STAT:QUES:ENAB 18  
Query Syntax  
Returned Parameters  
Related Commands  
STATus:QUEStionable:ENABle?  
<NR1>  
(Register value)  
STAT:QUES:EVEN?  
STAT:QUES:NTR|PTR Commands  
These commands allow the values of the Questionable NTR (Negative-Transition) and PTR (Positive-Transition) registers  
to be set or read. These registers serve as polarity filters between the Questionable Enable and Questionable Event registers  
to cause the following actions:  
When a bit of the Questionable NTR register is set to 1, then a 1-to-0 transition of the corresponding bit of the  
Questionable Condition register causes that bit in the Questionable Event register to be set.  
When a bit of the Questionable PTR register is set to 1, then a 0-to-I transition of the corresponding bit in the  
Questionable Condition register causes that bit in the Questionable Event register to be set.  
If the same bits in both NTR and PTR registers are set to 1, then any transition of that bit at the Questionable  
Condition register sets the corresponding bit in the Questionable Event register.  
If the same bits in both NTR and PTR registers are set to 0, then no transition of that bit at the Questionable  
Condition register can set the corresponding bit in the Questionable Event register.  
Note  
Setting a bit in the PTR or NTR filter can of itself generate positive or negative events in the  
corresponding Questionable Event register.  
Language Dictionary 45  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Command Syntax  
STATus:QUEStionable:NTRansition <NRf>  
STATus:QUEStionable:PTRansition <NRf>  
0 to 32727  
Parameters  
Suffix  
(None)  
Default Value  
Example  
Query Syntax  
0
STAT:QUES:NTR 16 STAT:QUES:PTR 512  
STATus:QUEStionable:NTRansition?  
STATus:QUEStionable:PTRansitiion?  
<NR1> (Register value)  
STAT:QUES:ENAB  
Returned Parameters  
Related Commands  
SYST:ERR?  
This query returns the next error number followed by its corresponding error message string from the remote programming  
error queue. The queue is a FIFO (first-in, first-out) buffer that stores errors as they occur. As it is read, each error is  
removed from the queue. When all errors have been read, the query returns 0,NO ERROR. If more errors are accumulated  
than the queue can hold, the last error in the queue is -350,TOO MANY ERRORS.  
Query Syntax  
Parameters  
SYSTem:ERRor?  
(None)  
Returned Parameters  
Example  
<NR1>,<SRD>  
SYST:ERR?  
SYST:VERS?  
This query returns the SCPI version number to which the power module complies. The returned value is of the form  
YYYY.V, where YYYY represents the year and V is the revision number for that year.  
Query Syntax  
Parameters  
SYSTem:VERSion?  
(none)  
Returned Parameters  
Example  
Related Commands  
<NR2>  
SYST:VERS?  
(None)  
Trigger Subsystem  
This subsystem controls the triggering of the power module. See "Chapter 5 - Synchronizing Power Module Output  
Changes" for an explanation of the Trigger Subsystem.  
Note  
The trigger subsystem must be enabled from the Initiate Subsystem or no triggering action will occur.  
TRIG  
When the trigger subsystem is enabled, TRIG generates an immediate trigger signal that bypasses any selected  
TRIG:SOUR and TRIG:DEL. The trigger will then:  
1.  
2.  
Initiate a pending level change as specified by CURR[:LEV]:TRIG or VOLT[:LEV]:TRIG.  
Initiate a pending level change as specified by CURR:MODE LIST or VOLT:MODE LIST and in accordance  
with LIST:STEP.  
3.  
Clear the WTG bit in the Status Operation Condition register.  
46 Language Dictionary  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Command Syntax  
Parameters  
TRIGger[:STARt][:IMMediate]  
(None)  
Examples  
TRIG TRIG: IMM  
Query Syntax  
Related Commands  
(None)  
ABOR CURR:MODE CURR:TRIG INIT *TRG  
VOLT:MODE VOLT:TRIG  
TRIG:DEL  
This command sets the time delay between the detection of an event on the specified trigger source and the start of any  
corresponding trigger action on the power module’s output.  
Command Syntax  
Parameters  
TRIGger[:STARt]:DELay <NRf+>  
0 to 65 | MIN | MAX  
Default Suffix  
S
*RST Value  
0
Examples  
Query Syntax  
Returned Parameters  
Related Commands  
TRIG:DEL .25 TRIG:DEL MAX  
TRIGger[:STARt]:DELay?  
<NR3>  
ABOR CURR:TRIG INIT *TRG TRIG[:IMM]  
VOLT:TRIG  
TRIG:LINK  
This command specifies which event conditions within the power module are linked to trigger source events when LINK is  
the parameter of the TRIG:SOUR command.  
Command Syntax  
Parameters  
TRIGger[:STARt]:LINK <CRD>  
(See Table 3-1)  
*RST Value  
OFF  
Examples  
Query Syntax  
Returned Parameters  
Related Commands  
TRIG:LINK "CC" TRIG:LINK "OPER"  
TRIGger[:STARt]:LINK?  
<CRD> (See Table 3-1)  
ABOR INIT *TRG TRIG[:IMM] TRIG:SOUR  
TRIG:SOUR  
This command selects the power module input trigger source as follows:  
BUS  
EXT  
*TRG or <GET> (Group Execute Trigger)  
Mainframe backplane Trigger In bus  
HOLD No trigger source except TRIG:IMM  
LINK  
TTLT  
Internal power module event as specified by TRIG:LINK  
Mainframe Trigger Out bus  
Command Syntax  
Parameters  
TRIGger[:STARt]:SOURce <CRD>  
BUS | EXT | HOLD | LINK | TTLT  
BUS  
*RST Value  
Examples  
Query Syntax  
Returned Parameters  
Related Commands  
TRIG: SOUR BUS TRIG: SOUR LINK  
TRIGger[:STARt]:SOURce?  
BUS | EXT | HOLD | TTLT  
ABOR CURR:TRIG INIT OUTP:TTLT  
VOLT:TRIG  
Language Dictionary 47  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Voltage Subsystem  
This subsystem programs the output voltage of the power module.  
VOLT  
This command directly programs the immediate voltage level of the power module. The immediate level is the voltage  
applied at the output terminals. This command is always active, even when the voltage subsystem is in the list mode (see  
VOLT:MODE).  
Command Syntax  
Parameters  
[SOURce]:VOLTage[:LEVel] [:IMMediate][:AMPLitude] <NRf+>  
Table 3-2  
Default Suffix  
*RST Value  
Examples  
V
Table 3-2  
VOLT 2500 MV VOLT:LEV 2.5  
Query Syntax  
[SOURce]:VOLTage[:LEVel] [:IMMediate][:AMPlitude]?  
[SOURce]:VOLTage[:LEVel] [:IMMediate][:AMPLitude]? MAX  
[SOURce]:VOLTage[:LEVel] [:IMMediate][:AMPLitude]? MIN  
<NR3> VOLT? returns the presently programmed immediate voltage  
level.  
VOLT? MAX and VOLT? MIN return the maximum and minimum  
programmable immediate voltage levels.  
*SAV *RCL *RST  
Returned Parameters  
Related Commands  
VOLT:MODE  
This command enables or disables list subsystem control over the power module output voltage. When programmed with  
FIX, this command prevents the output voltage from being controlled by the sequencing of points specified by  
LIST:VOLT. If the LIST parameter is used, then the output voltage may be changed by the subsequent execution of a list.  
However, the list mode does not prevent the output voltage from being set by VOLT[:IMM] and *RCL.  
Note  
VOLT:MODE LIST is an implied ABORT command.  
Command Syntax  
Parameters  
[SOURce]:VOLTage:MODE <CRD>  
FIXed | LIST  
*RST Value  
FIX  
Examples  
Query Syntax  
Returned Parameters  
Related Commands  
VOLT:MODE LIST VOLT:MODE FIX  
[SOURce]:VOLTage:MODE?  
FIX | LIST  
VOLT:LIST *RCL  
VOLT:PROT  
This command sets the overvoltage protection (OVP) level of the power module. If the output voltage exceeds the OVP  
level, then the power module output is disabled and the Questionable Condition status register OV bit is set (see "Chapter 4  
- Status Reporting"). An overvoltage condition can be cleared with the OUTP:PROT:CLE command after the condition  
that caused the OVP trip is removed. The OVP always trips with zero delay and is unaffected by the OUTP:PROT:DEL  
command.  
Command Syntax  
Parameters  
[SOURce]:VOLTage:PROTection [:LEVel] <NRf+>  
See Table 3-2  
Default Suffix  
*RST Value  
Examples  
V
MAX  
VOLT:PROT 2.5 VOLT:PROT:LEV MAX  
48 Language Dictionary  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Query Syntax  
Returned Parameters  
Related Commands  
[SOURce]:VOLTage:PROTection [:LEVel]?  
[SOURce]:VOLTage:PROTection [:LEVel]? MIN  
[SOURce]:VOLTage:PROTection [:LEVel]? MAX  
<NR3> VOLT:PROT? returns presently programmed OVP level.  
VOLT:PROT? MAX and VOLT:PROT? MIN return the  
maximum and minimum programmable OVP levels.  
OUTP:PROT:CLE *RST *SAV *RCL  
VOLT:SENS:SOUR?  
This command reads the state of the power module output connector remote sense switch. The INTernal parameter  
corresponds to the LOCAL position of the switch. (See the power module User’s Guide for more information about this  
switch.) VOLT:SENS:SOUR is an alias for the SCPI VOLT:ALC:SOUR command.  
Query Syntax [SOURce]:VOLTage:SENSe:SOURce?  
Example VOLT:SENS:SOUR?  
Returned Parameters EXTernal | INTernal  
Related Commands VOLT:ALC:SOUR?  
VOLT:TRIG  
This command programs the pending triggered voltage level of the power module. The pending triggered voltage level is a  
stored value that is transferred to the output terminals when a trigger occurs. A pending triggered level is unaffected by  
subsequent VOLT:LEV[:IMM] commands and remains in effect until the trigger subsystem receives a trigger or an  
ABORt command is given. In order for VOLT:TRIG to be executed, the trigger subsystem must be initiated (see  
INITiate).  
Command Syntax  
Parameters  
[SOURce]:VOLTage[:LEVel]:TRIGgered [:AMPLitude] <NRf+>  
See Table 3-2  
Default Suffix  
*RST Value  
Examples  
V
See Table 3-2  
VOLT:TRIG 1200 MV VOLT:LEV:TRIG 1.2  
Query Syntax  
[SOURce]:VOLTage[:LEVel]:TRIGgered [:AMPLitude]?  
[SOURce]:VOLTage[LEVel]:TRIGgered [:AMPLitude]? MAX  
[SOURce]:VOLTage[:LEVel]:TRIGgered [:AMPLitude]? MIN  
<NR3> VOLT:TRIG? returns the presently programmed current  
level. If the TRIG level is not programmed, the IMM level is returned.  
VOLT:TRIG? MAX and VOLT:TRIG? MIN return the  
maximum and minimum programmable triggered voltage levels.  
ABOR VOLT[:IMM] VOLT:MODE INIT *RST  
Returned Parameters  
Related Commands  
Language Dictionary 49  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table 3-1. Link Parameter List  
Valid for  
Parameter  
CC  
True Event Condition  
Constant current event bit1  
Constant voltage event bit1  
Standard event summary bit1  
List sequence complete pulse2  
Operation complete bit1  
OUTP:DFI:LINK  
OUTP:DFI:LINK  
OUTP:DFI:LINK  
OUTP:TTLT:LINK  
OUTP:TTLT:LINK  
OUTP:TTLT:LINK  
OUTP:TTLT:LINK  
OUTP:TTLT:LINK  
OUTP:TTLT:LINK  
OUTP:TTLT:LINK  
OUTP:TTLT:LINK  
OUTP:TTLT:LINK  
OUTP:TTLT:LINK  
OUTP:TTLT:LINK  
OUTP:TTLT:LINK  
OUTP:TTLT:LINK  
OUTP:TTLT:LINK  
OUTP:TTLT:LINK  
TRIG[:STAR]:LINK  
TRIG[:STAR]:LINK  
TRIG[:STAR]:LINK  
TRIG[:STAR]:LINK  
TRIG[:STAR]:LINK  
TRIG[:STAR]:LINK  
TRIG[:STAR]:LINK  
TRIG[:STAR]:LINK  
TRIG[:STAR]:LINK  
TRIG[:STAR]:LINK  
TRIG[:STAR]:LINK  
TRIG[:STAR]:LINK  
TRIG[:STAR]:LINK  
TRIG[:STAR]:LINK  
TRIG[:STAR]:LINK  
CV  
ESB  
LSC  
OPC  
OPER  
QUES  
MAV  
OFF  
RQS  
RTG  
STC  
STS  
SUM3  
TDC  
OUTP:DFI:LINK  
OUTP:DFI:LINK  
OUTP:DFI:LINK  
OUTP:DFI:LINK  
OUTP:DFI:LINK  
OUTP:DFI:LINK  
Operation summary bit1  
Questionable summary bit1  
Message available summary bit1  
No linked event condition  
Request service summary bit1  
Received a trigger bit2  
List step completed pulse2  
List step started pulse2  
OPER or QUES or ESB bit1  
Trigger delay complete pulse 2  
OUTP:DFI:LINK  
1 See “Chapter 4 - Status Reporting” 2 See “Chapter 5 - Synchronizing Power Module Output Changes”.  
Table 3-2. Power Module Programming Parameters  
Agilent Model  
Parameter  
66101A  
66102A  
66103A  
66104A  
66105A  
66106A  
Output Programming Range  
(maximum programmable values):  
Voltage:  
8.190 V  
16.380 A  
8.8 V  
20.475 V  
7.678 A  
22.0 V  
35.831 V  
4.607 A  
38.5 V  
61.425 V  
2.559 A  
66.0 V  
122.85 V  
1.280 A  
132.0 V  
204.75 V  
0.768 A  
220.0 V  
Current:  
OV Protection:  
Average Resolution  
Voltage:  
2.4 mV  
4.6 mA  
50 mV  
5.9 mV  
2.3 mA  
120 mV  
10.4 mV  
1.4 mA  
200 mV  
18.0 mV  
0.75 mA  
375 mV  
36.0 mV  
0.38 mA  
750 mV  
60.0 mV  
0.23 mA  
1.25 V  
Current:  
OV Protection:  
*RST State Values1  
Voltage:  
0
257 mA  
8.8 V  
120 mA  
22.0 V  
72 mA  
38.5 V  
40 mA  
66.0 V  
20 mA  
132.0 V  
12 mA  
220.0 V  
Current:  
OV Protection:  
1 These also are the power-on reset values when the factory default parameters are left in effect (see "Chapter 2  
Installation” in the power module User’s Guide).  
Nonvolatile Memory Locations:  
Volatile Memory Locations:  
5; (0 through 4)  
5; (5 through 9)  
:
50 Language Dictionary  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4
Status Reporting  
Power Module Status Structure  
Figure 4-1 shows the status register structure of the power module. The Standard Event, Status Byte, and Service Request  
Enable registers and the Output Queue perform standard GPIB functions as defined in the IEEE 488.2 Standard Digital  
Interface for Programmable Instrumentation. The Operation Status and Questionable Status registers implement status  
functions specific to the power module.  
Status Register Bit Configuration  
Table 4-2 and Figure 4-1 show the bit configuration of each status register.  
Operation Status Group  
Register Functions  
The Operation Status registers record signals that occur during normal operation. The group consists of the following  
registers:  
A Condition register that holds real-time status of the circuits being monitored. It is a read-only register.  
A PTR/NTR (positive transition/negative transition) Filter that functions as described under  
STAT:OPER:NTR|PTR COMMANDS in "Chapter 3 - Language Dictionary". This is a read/write register.  
An Event register that latches any condition that is passed through the PTR or NTR filters. Reading the Event  
register clears it.  
An Enable register that functions as described under STAT:OPER:ENAB in "Chapter 3 -Language  
Dictionary". This is a read/write register.  
The outputs of the Operation Status group are logically-ORed into the OPER(ation) summary bit (7) of the Status Byte  
register.  
Register Commands  
Commands that access this group are derived from the STAT:OPER commands described in "Chapter 3 - Language  
Dictionary" and summarized in Table 4-1.  
Table 4-1. Status Operation Commands  
Register  
Command  
Query  
Cleared By  
Condition  
(None)  
STAT:OPER:COND?  
Cannot be cleared  
PTR Filter  
NTR Filter  
Event  
STAT:OPER:PTR <NRf>  
STAT:OPER:NTR <NRf>  
(None)  
STAT:OPER:PTR?  
STAT:OPER:NTR?  
STAT:OPER:EVEN?  
STAT:OPER:ENAB?  
Programming 0 or  
STAT:PRES  
Programming 0  
Reading or *CLS  
Programming 0  
Enable  
STAT:OPER:ENAB <NRf>  
Status Reporting 51  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table 4-2. Bit Configurations of Status Registers  
Bit Signal  
Meaning  
Bit Signal  
Meaning  
Operation Status Group  
Standard Event Status Group  
0
CAL  
The interface is computing new  
calibration constants  
0
OPC  
Operation complete  
5
8
WTG  
CV  
The interface is waiting for a trigger  
The power module is in constant  
voltage mode  
2
3
QYE  
DDE  
Query error  
Device-dependent error  
10  
12  
CC  
The power module is in constant  
current mode  
The list step is active (dwelling)  
4
EXE  
Execution error  
DWE  
5
7
CME  
PON  
Command error  
Power on  
Questionable Status Group  
Status Byte and Service Request  
Enable Registers  
0
OV  
OC  
OT  
RI  
The power module overvoltage  
protection circuit has tripped  
The power module overcurrent  
protection circuit has tripped  
The power module has an  
overtemperature condition  
The power module remote inhibit state  
is active  
3
4
5
6
7
QUES  
MAV  
ESB  
Questionable status summary bit  
1
Message Available summary bit  
Event Status summary bit  
4
9
MSS  
RQS  
OPER  
Master Status summary bit  
Request Service bit  
Operation status summary bit  
10  
UNR  
The power module output is  
unregulated  
Questionable Status Group  
Register Functions  
The Questionable Status registers record signals that indicate abnormal operation of the power module. As shown in Figure  
4-1, the group consists of the same type of registers as the Status Operation group. The outputs of the Questionable Status  
group are logically-ORed into the QUES(tionable) summary bit (3) of the Status Byte register.  
Register Commands  
Programming for this group is derived from the STAT:QUES commands described in "Chapter 3 - Language Dictionary"  
and summarized in Table 4-3.  
Table 4-3. Status :Questionable Commands  
Register  
Command  
Query  
Cleared By  
Condition  
(None)  
STAT:QUES:COND?  
Cannot be cleared  
PTR Filter  
NTR Filter  
Event  
STAT:QUES:PTR <NRf>  
STAT:QUES:NTR <NRf>  
(None)  
STAT:QUES:PTR?  
STAT:QUES:NTR?  
STAT:QUES:EVEN?  
STAT:QUES:ENAB?  
Programming 0  
Programming 0 or  
STAT:PRES  
Reading or *CLS  
Enable  
STAT:QUES:ENAB <NRf>  
Programming 0  
52 Status Reporting  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Figure 4-1. Power Module Status Model  
Standard Event Status Group  
Register Functions  
This group consists of an Event register and an Enable register that are programmed by COMMON commands. The  
Standard Event register latches events relating to interface communication status (see Table 4-2). It is a read-only register  
that is cleared when read. The Standard Event Enable register functions similarly to the enable registers of the Operation  
and Questionable status groups.  
Register Commands  
The common *ESE command programs specific bits in the Standard Event Status Enable register. Because the power  
module implements *PSC, the register is cleared at power on if *PSC = 1.  
*ESR? reads the Standard Event Status Event register. Reading the register clears it.  
Status Reporting 53  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Status Byte Register  
This register summarizes the information from all other status groups as defined in the IEEE 488.2 Standard Digital  
Interface for Programmable Instrumentation standard. The bit configuration is shown in Table 4-2. The register can be  
read either by a serial poll or by *STB?. Both methods return the same data, except for bit 6. Sending *STB? returns MSS  
in bit 6, while polling returns RQS in bit 6.  
The RQS Bit  
Whenever the power module requests service, it sets the SRQ interrupt line true and latches RQS into bit 6 of the Status  
Byte register. When the controller services the interrupt, RQS is cleared inside the register and returned in bit position 6 of  
the response. The remaining bits of the Status Byte register are not disturbed.  
The MSS Bit  
This is a real-time (unlatched) summary of all Status Byte register bits that are enabled by the Service Request Enable  
register. MSS is set whenever the power module has at least one reason (and possible more) for requesting service. Sending  
*STB? reads the MSS in bit position 6 of the response. No bits of the Status Byte register are cleared by reading it.  
Determining the Cause of a Service Interrupt  
You can determine the reason for an SRQ by the following actions:  
Use a serial poll or the *STB? query to determine which summary bits are active.  
Read the corresponding Event register for each summary bit to determine which events caused the summary  
bit to be set. When an Event register is read, it is cleared. This also clears the corresponding summary bit.  
The interrupt will recur until the specific condition that caused the each event is removed. If this is not  
possible, the event may be disabled by programming the corresponding bit of the status group Enable register or  
NTR|PTR filter. A faster way to prevent the interrupt is to disable the service request by programming the  
appropriate bit of the Service Request Enable register.  
Output Queue  
The Output Queue is a first-in, first-out (FIFO) data register that stores power module-to-controller messages until the  
controller reads them. Whenever the queue holds one or more bytes, it sets the MAV bit (4) of the Status Byte register. If  
too many unread error messages are accumulated in the queue, a system error message is generated (see "Chapter 6 - Error  
Messages"). The Output Queue is cleared at power on and by *CLS.  
Location Of Event Handles  
"Event handles" are signals within the interface that can be used for triggers, for a Trigger Out signal, or for a DFI signal.  
Those event handles derived from signals in the Status Subsystem are shown as circled numbers in Figure 4-1. Other event  
handles are described in "Chapter 5 - Synchronizing Power Module Output Changes".  
54 Status Reporting  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Initial Conditions At Power On  
Status Registers  
When the power module is turned on, a sequence of commands initializes the status registers. Table 4-4 shows the register  
states and corresponding power-on commands for the factory-default *RST power-on state. If the module power-on function  
switch is set to 0, then the power-on state is determined by the parameters stored in location 0 (see Chapter 4 of the User’s  
Guide).  
Table 4-4. Default Power On Register States  
Register  
Condition  
Caused By  
STAT:PRE  
Operation PTR; Questionable PTR  
Operation NTR; Questionable NTR  
Operation Event; Questionable Event  
Operation Enable; Questionable Enable  
Standard Event Status Enable  
Status Byte  
All bits = 1  
All bits = 0  
All bits = 0  
All bits = 0  
All bits = 01  
All bits = 0  
All bits = 01  
Cleared  
STAT:PRE  
*CLS  
STAT:PRE  
*ESE 0  
*CLS  
Status Request Enable  
*SRE 0  
*CLS  
Output Queue  
1If PSC=1. If PSC = 0, then the last previous state before turn on is recalled. The value of PSC is  
stored in nonvolatile memory.  
The PON (Power On) Bit  
The PON bit in the Standard Event register is set whenever the power module is turned on. The most common use for PON  
is to generate an SRQ at power on following an unexpected loss of power. To do this, bit 7 of the Standard Event Enable  
register must be set so that a power-on event registers in the ESB (Standard Event Summary Bit). Also, bit 5 of the Service  
Request Enable register must be set to permit an SRQ to be generated. The commands to accomplish these two conditions  
are:  
*ESE 128  
*SRE 32  
If *PSC is programmed to 0, the contents of the Standard Event Enable and Service Request Enable registers are saved in  
nonvolatile memory and recalled at power on. This allows a PON event to generate SRQ at power on. Programming *PSC  
to 1 prevents these registers from being saved and they are cleared at power on. This prevents a PON event from generating  
SRQ at power on.  
Examples  
Note  
These examples are generic SCPI commands. See "Chapter 2 - Programming Introduction" for  
information about encoding the commands as language strings.  
Servicing an Operation Status Mode Event  
This example assumes you want a service request generated whenever the power module switches to the CC (constant  
current) mode. From Figure 4-1, note that the required path is for a condition at bit 10 (CC) of the Operation Status register  
to set bit 6 (RQS) of the Status Byte register. The required register programming is as follows:  
Status Reporting 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table 4-5. Generating RQS from the CC Event  
Register  
Operation PTR  
Command  
STAT:OPER:PTR 1024  
Comment  
Allows a positive transition at the CC input (bit 10) to be  
latched into the Status Event register.1  
Allows the latched CC event to be summed into the OPER  
summary bit.  
Enables the OPER summary bit from the Status Byte register to  
generate RQS.  
When you service the request, read the event register to  
determine that bit 10 (CC) is set and to clear the register for the  
next event.  
Operation Enable  
STAT:OPER:ENAB 1024  
*SRE 128  
Service Request Enable  
Operation Condition  
STAT:OPER:EVEN?  
1All bits of the PTR registers bits are set to 1 at power on.  
Adding More Operation Events  
To add CV (constant voltage) and DWE (dwelling) events to this example, it is only necessary to add the decimal values for  
bit 8 (value 64) and bit 12 (value 4096) to the programming commands of the Operation Status group. The commands to do  
this are:  
STAT:OPER:PTR 5376;ENAB 5376  
It is not necessary to change any other registers, since the programming for the operation summary bit (OPER) path has  
already been done.  
Servicing Questionable Status Events  
To add OC (overcurrent) and OT (overtemperature) events to this example, program Questionable Status group bits  
1 and 4.  
STAT:QUES:PTR 18;ENAB 18  
Next, you must program the Service Request Enable register to recognize both the questionable (QUES) and the operational  
(OPER) summary bits.  
*SRE 136  
Now when there is a service request, read back both the operational and the questionable event registers.  
STAT:OPER:EVEN?;QUES:EVEN?  
Monitoring Both Phases of a Status Transition  
You can monitor a status signal for both its positive and negative transitions. For example, to generate RQS when the  
power module either enters the CC (constant current) condition or leaves that condition, program the Operational Status  
PTR/NTR filter as follows:  
STAT:OPER:PTR 1024;NTR 1024  
STAT:OPER:ENAB 1024;*SRE 128  
The PTR filter will cause the OPER summary bit to set RQS when CC occurs. Then the controller subsequently reads the  
event register (STAT: OPER: EVEN?), the register is cleared. When CC subsequently goes false, the NTR filter causes the  
OPER summary bit to again set RQS.  
56 Status Reporting  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5
Synchronizing Power Module Output Changes  
Introduction  
If you use only the VOLT [: LEV] : TRIG and/or CURR [: LEV] : TRIG commands to trigger output changes, you do not need the  
information in this chapter. This chapter gives supplemental information on how you can synchronize power module output  
changes to internal or external events. The output changes can be:  
A change in output voltage level.  
A change in output current level.  
The start of an internally-paced list of output voltage or current levels.  
A step to the next level in a list of output voltage or current levels.  
A change in the output state (on or off).  
The event to which the output change is synchronized can be any of the following:  
A command from the controller.  
A GPIB bus command.  
An event that occurs within the power module.  
An event that occurs within another power module.  
An external signal at the mainframe trigger input.  
An external signal at the mainframe fault inhibit input (INH).  
The output synchronization is implemented by the following power module functions:  
Trigger subsystem.  
List subsystem.  
Remote inhibit (RI) subsystem.  
Discrete fault indicator (DFI) subsystem.  
Trigger Subsystem  
Two simplified models of the trigger subsystem are presented. The first model shows how the trigger subsystem functions  
during fixed-mode output. This mode occurs when VOLT:MODE FIX or CURR:MODE FIX is in effect (see "Chapter 3 -  
Language Dictionary"). In this mode of output control, each triggered output voltage or current value is explicitly specified  
by a triggered-level command (for example, VOLT: TRIG 20 or CURR: TRIG 1.55). The trigger then causes the output to  
change to this pending triggered level.  
The second model shows the difference in trigger subsystem operation during list-mode output. This mode occurs when the  
VOLT:MODE LIST or CURR:MODE LIST command is programmed (see "List Subsystem" further in this chapter for an  
explanation of lists). In this mode of output operation, the triggered output voltage or current levels are specified within a  
list and the trigger controls the sequencing through the values in the list.  
Model of Fixed-Mode Trigger Operation  
Figure 5-lA is a simplified model of trigger subsystem operation when the power module is programmed for fixed-mode  
output. The rectangular boxes represent states. The arrows show the transitions between states. These are labeled with the  
input or event that causes the transition to occur.  
Synchronizing Power Module Output Changes 57  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Idle State  
When the power module is turned on, the trigger subsystem is in the idle state. In this state, the trigger subsystem ignores  
all triggers. When the trigger action has been completed, the trigger subsystem returns to this state. It also returns to the  
Idle state if the ABORt command or an implied ABORt command (*RST, *RCL, or any LIST) is sent.  
Initiated State  
The INITiate command moves the trigger subsystem from the Idle state to the Initiated State. This enables the power module  
to receive triggers. The source of the trigger is selected with the TRIGger:SOURce command (see "Chapter 3 - Language  
Dictionary"). When in the Initiated state, the power module responds to events on the selected trigger source by transferring  
to the Delaying state. As shown in Figure 5-lA there is another trigger signal that is not subject to TRIG: SOUR control. This  
is the TRIGger:IMMediate command. If the trigger subsystem is in the Initiated state, this command generates a trigger that  
transfers the trigger subsystem directly to the Output Change state, bypassing the Delaying state.  
Delaying State  
When a trigger event occurs on the selected trigger source, the trigger subsystem transfers to the Delaying state. In this  
state, the subsystem waits for the interval specified by the TRIGger:DELay command before moving to the next state. As  
shown in Figure 5-1A, a TRIGger: IMMediate command will bypass any programmed delay and cause an immediate transition  
to the Output Change state.  
Figure 5-1. Simplified Models of Trigger Modes  
58 Synchronizing Power Module Output Changes  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Output Change State  
When the trigger subsystem enters the Output Change state, the output voltage and current are set to the pending levels  
programmed by the VOLTage:TRIGgered and CURRent:TRIGgered commands. Once this occurs, the existing triggered levels  
are cleared and must be reprogrammed. If no triggered levels are programmed, then the trigger has no effect on the output  
levels. When the triggered actions are completed, the trigger subsystem returns to the Idle state.  
Model of List Mode Trigger Operation  
Figure 5-1B is a simplified model of trigger subsystem operation when power module is programmed for list-mode output.  
Operation in the Idle, Initiated, and Delaying states are identical to that described under fixed-mode operation.  
Output Change State  
When the trigger subsystem enters the Output Change state in List mode, the output voltage and/or current is set to the next  
value (point) in the programmed list. The trigger subsystem then transfers to the next (Dwelling) state and increments the  
list to the next point. If there are no more points in the list, the subsystem resets the list to the first point. This completes  
the list, unless LIST: COUNt is programmed to greater than one. In that case, the list is not completed until it has repeated the  
list sequence the number of times specified by the count.  
Dwelling State  
Each voltage and current list point has an associated dwell interval specified by the LIST:DWELL command. After the new  
output value is established, the trigger system pauses for the programmed dwell interval. During this dwell interval, trigger  
events are ignored and only an ABORt (or implied abort) command can transfer the subsystem out of the Dwelling state.  
At the end of the dwell interval, the transition to the next state depends on whether or not the list has completed its  
sequencing and on how the LIST: STEP command has been programmed.  
If the list is completed, the trigger subsystem returns to the Idle state.  
If the list is not completed, then the subsystem reacts as follows:  
If LIST: STEP ONCE has been programmed, the trigger subsystem returns to the Idle state.  
If LIST: STEP AUTO has been programmed, the trigger subsystem returns to the Output Change state and  
immediately executes the next list point.  
The INITiate:CONTinuous Command  
In the above descriptions of the trigger subsystem models, the INITiate: IMMediate command was used to move from the Idle  
to the Initiated state. In some applications, it may be desirable to have the subsystem return directly to the Initiated state  
after a trigger action has completed. Programming INITiate:CONTinuous ON does this by bypassing the Idle state. If the  
ABORt command is given while INIT: CONT is ON, the trigger subsystem transfers to the Idle state but immediately exits to  
the Initiated state.  
Trigger Status and Event Signals  
Some transitions of the trigger subsystem provide inputs to the status subsystem. Others are defined as "event handles",  
which are selectable trigger sources by way of parameters in link commands TRIGger:LINK, OUTPut:DFI:LINK and  
OUTPut:TTLT:LINK (see Table 3-1). Table 5-1 summarizes these signals.  
Synchronizing Power Module Output Changes 59  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table 5-1. Trigger Subsystem Status and Event Signals  
Signal  
DWE  
Type  
Status Bit  
Description  
Dwelling. True only during the dwelling state. DWE can be monitored at the  
Operation Status register (see "Chapter 4 - Status Reporting").  
LSC  
Event Handle List Sequence Complete. Occurs upon exit from the Dwelling state after the  
last programmed list point has been executed. If LIST: COUNt is greater than  
1, LSC occurs once for each count until the list is done.  
RTG  
TDC  
STC  
STS  
Event Handle Received a trigger. Occurs upon exit from the Initiated state.  
Event Handle Trigger delay complete. Occurs upon exit from the Delaying state.  
Event Handle Step Completed. Occurs upon exit from the Dwelling state.  
Event Handle Step Started. Occurs upon transition into the Dwelling state.  
WTG  
Status Bit  
Waiting for trigger. True only when the trigger subsystem is in either the  
Initiated or the Delaying state. WTG can be monitored at the Operation Status  
register (see "Chapter 4 - Status Reporting”).  
Trigger In and Trigger Out  
The mainframe has two bnc connectors labeled Trigger In and Trigger Out. Figure 5-2 shows the model for these signals,  
which are applied to all power modules in the mainframe (see “TrigIn/TrigOut Characteristics” in Chapter 1 of Agilent  
66000A Installation Guide for electrical parameters). Trigger In and Trigger Out are electrically isolated at each power  
module from the mainframe chassis reference ground.  
Trigger In  
Trigger In is a TTL level input that can be selected as a trigger source for each module. Modules recognize a Trigger In  
signal on its falling edge. Trigger In is selected as a trigger source with the EXTernal parameter. For example:  
TRIGger:SOURce EXT  
OUTPut:TTLT:SOURce EXT  
Trigger Out  
The Trigger Out signal is a 20-microsecond, negative-true TTL pulse. This pulse can be driven by each power module by  
programming the OUTPut:TTLTrg commands. Each module can also select Trigger Out as a trigger source by programming  
the SCPI TRIGger:SOURce TTLT command (see "Chapter 3 - Language Dictionary" for details of these commands).  
To select the Trigger In connector as a trigger source, use TRIG: SOUR EXT  
To apply a trigger to the Trigger Out connector, use OUTP:TTLT ON. You must also select the source  
(OUTP:TTLT:SOUR).  
Figure 5-2. TTLT Trigger Model  
60 Synchronizing Power Module Output Changes  
Download from Www.Somanuals.com. All Manuals Search And Download.  
List Subsystem  
The List Subsystem commands allow you to program a sequence of voltage and/or current values that will be applied to the  
power module output when it is the list mode (VOLTage:MODE LIST or CURRent:MODE LIST). Up to 20 voltage and current  
values, with 20 associated time intervals (dwells), may be programmed. By using lists, you can program a complex  
sequence of power module outputs with minimal interaction between the controller and the power module. Lists allow you  
to time output changes more precisely or to better synchronize them (using triggers) with asynchronous events.  
Basic Steps of List Sequencing  
You can program the number of output levels (or points) in the list, the time interval that each level is maintained, the  
number of times that the list will be executed, and how the levels change in response to triggers. This is a synopsis of the  
list commands:  
List Function  
Command  
Enable the voltage list function  
Enable the current list function  
VOLT:MODE LIST  
CURR:MODE LIST  
LIST:VOLT <NRf+>  
LIST:CURR <NRf+>  
LIST:DWEL <NRf+>  
LIST:COUN <NRf+>  
LIST:STEP AUTO|ONCE  
Specify the voltage output levels (points)  
Specify the current output levels (points)  
Specify the time duration of each output level  
Specify the times the list is repeated  
Select the list response to a trigger  
Programming the List Output Levels  
1. Enable the specific output to be controlled by the list. For example,  
VOLT:MODE LIST  
CURR:MODE LIST  
2. Program the desired output levels or points. The order of the points determines the order in which the output levels will  
occur. To sequence the voltage through values of 1, 1.5, 3, 1.5, and 1 volts, program:  
LIST:VOLT 1,1.5,3.0,1.5,1  
You can specify lists for both voltage and current. For example:  
LIST:VOLT 1,2,5,6,8  
LIST:CURR 10,5,2,1.67,1.25  
Both lists must have the same number of points. The exception is if a list has only a single point. In this case, the single-  
point list is treated as if it has the same number of points as the other list with each point equal to the programmed value.  
For example, if you send:  
LIST:VOLT 1,2,5,6,8;CURR 1  
then the power module will respond as if the two lists were:  
LIST:VOLT 1,2,5,6,8  
LIST:CURR 1,1,1,1,1  
Note  
Execution of a list will be aborted if an ABORt command or an implied ABORt command (another list  
command, the *RST command or the *RCL command) is sent.  
Programming List Intervals  
The dwell time is the interval that the output remains at the programmed value. The time unit is seconds. The following  
command specifies five dwell intervals:  
LIST:DWEL 1,1.5,3,1.5,.5  
Synchronizing Power Module Output Changes 61  
Download from Www.Somanuals.com. All Manuals Search And Download.  
The number of dwell points must equal the number of output points:  
LIST:VOLT 3.0,3.25,3.5,3.75  
LIST:DWEL 10,10,25,40  
The only exception is for a dwell list with one value, which gives the same interval to all the points in the corresponding  
voltage or current list.  
Note  
Sending a VOLT [: LEV: IMM] or CURR [: LEV: IMM] command during an interval will override the list  
output value for that interval. When the next interval begins, the output will be determined by the list  
value for that interval.  
Automatically Repeating a List  
You can repeat a list by entering a LIST: COUNt parameter. The parameter determines how many times a list is executed or  
sequenced. Enter an integer or enter the value INF to make the list repeat indefinitely. For example, to make the current list  
2,3,12,15 repeat 5 times, send:  
LIST:CURR 2,3,12,15  
LIST:COUN 5  
The LIST: COUNt parameter is stored by *SAV and restored by *RCL. The GPIB *RST value is 1.  
Triggering a List  
No list will execute without a trigger. How the list responds to a trigger depends on how you program the LIST: STEP AUTO  
| ONCE command. The method you use will depend upon whether you want the list to be paced by dwell intervals or by  
triggers.  
Dwell-Paced Lists  
For a closely controlled sequence of output levels, you can use a dwell-paced list. Each list output point remains in effect  
for the dwell time associated with that point. When the dwell time expires, the output immediately changes to the next point  
in the list.  
For dwell pacing, program LIST: STEP to AUTO (see Figure 5-3-A). The dwell-paced list requires only a single trigger to  
start the list. The trigger subsystem remains in the dwelling state until the list is completed. If LIST: COUN is greater than 1,  
the entire list is repeated until the count has been satisfied (see Figure 5-1-B).  
Trigger-Paced Lists  
If you need the output to closely follow asynchronous events, then a trigger-paced list is more appropriate. Program LIST:  
STEP to ONCE (see Figure 5-3-B). Now expiration of a dwell interval returns the trigger subsystem to the Initiated state.  
The subsystem then waits for a trigger to start the next dwell interval. During this time, the power module output remains at  
the level set by the last executed point in the list.  
Note  
If the subsystem is not in the dwelling state, a TRIGger [: IMMediate] command will sequence the next point  
in the list.  
62 Synchronizing Power Module Output Changes  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Figure 5-3. Timing diagrams of LIST:STEP Operation  
Synchronizing Power Module Output Changes 63  
Download from Www.Somanuals.com. All Manuals Search And Download.  
DFI (Discrete Fault Indicator) Subsystem  
Whenever a fault is detected in the power module, it is capable of generating a low-true TTL signal at the mainframe FLT  
jack for communication with external devices (see “INH/FLT Characteristics” in Chapter 1 of the Agilent 66000A  
Installation Guide for the electrical parameters). The source for the DFI signal can be any of the parameters of the  
OUTPut:DFI:LINK command (see Table 3-1). The SUM3 link parameter allows any combination of Questionable, Operation,  
or Event status bits to generate the DFI signal. The GPIB *RST command sets the link parameter to SUM3.  
RI (Remote Inhibit) Subsystem  
Each power module is connected to the mainframe INH jack via a function selector switch. (See Chapter 2 of the Operating  
Guide for details concerning this switch.) When the switch is set to enable the RI function, a low-true TTL signal at the INH  
input will shut down the power module. This generates an RI status bit at the Questionable Status register (see "Chapter 4 -  
Status Reporting"). By programming the status subsystem, you may use RI to generate a service request (SRQ) to the  
controller and/or to create a DFI output at the mainframe FLT jack. By using RI/DFI in this way, you can chain the power  
modules to create a serial shutdown in response to the INH input.  
SCPI Command Completion  
SCPI commands sent to the power module are processed either sequentially or in parallel. Sequential commands finish  
execution before a subsequent command begins. A parallel command can begin execution while a preexisting command is  
still executing (overlapping commands). Commands that affect list and trigger actions are among the parallel commands.  
There *WAI, *OPC, and *OPC? common commands provide different ways of indicating when all transmitted commands,  
including any parallel ones, have completed their operations. The syntax and parameters for these commands are described  
in "Chapter 3 - Language Dictionary". Some practical considerations for using these commands are as follows:  
*WAI  
This prevents the power module from processing subsequent commands until all pending operations are  
completed. If something prevents completion of an existing operation, *WAI can place the module and the  
controller in a "hang-up" condition.  
*OPC?  
*OPC  
This places a 1 in the Output Queue when all pending operations have completed. Because it requires  
your program to read the returned value from the queue before executing the next program statement,  
*OPC? could prevent subsequent commands from being executed.  
This sets the OPC status bit when all pending operations have completed. Since your program can read  
this status bit on an interrupt basis, *OPC allows subsequent commands to be executed.  
The trigger subsystem must be in the Idle state in order for the status OPC bit to be true. Therefore, as far as triggers and  
lists are concerned, OPC is false whenever the trigger subsystem is in the Initiated state. However, OPC is also false if there  
are any commands pending within any other subsystems. For example, if you sendCURR: TRIG 1 . 5 after a VOLT:LIST  
command, completion of the CURR:TRIG command will not set OPC if the list command is still executing.  
Note  
For a detailed discussion of *WAI, *OPC and *OPC?, see "Device/Controller Synchronization Techniques"  
in ANSI/IEEE Std. 488.2-1987.  
64 Synchronizing Power Module Output Changes  
Download from Www.Somanuals.com. All Manuals Search And Download.  
6
Error Messages  
Power Module Hardware Error Messages  
Front panel error messages resulting from selftest errors or runtime failures are described in the power module User’s  
Guide.  
System Error Messages  
System error messages are read back via the SYST:ERR? query. The error number is the value placed in the power module  
error queue. SYST:ERR? returns the error number into a variable and combines the number and the error message into a  
string. Table 6-1 lists the system errors that are associated with SCPI syntax errors and interface problems. Information  
inside the brackets is not part of the standard error message, but is included for clarification. When system errors occur, the  
Standard Event Status register (see "Chapter 4 - Status Reporting") records them as follows:  
Standard Event Status Register Error Bits  
Bit Set  
Error Code  
-100 thru -199  
-200 thru -299  
Error Type  
Command  
Execution  
Bit Set  
Error Code  
-300 thru -399  
-400 thru -499  
Error Type  
Device-dependent  
Query  
5
4
3
2
Table 6-1. Summary of System Error Messages  
Error String [Description/Explanation/Examples]  
Error  
Number  
-100  
-101  
-102  
-103  
-104  
-105  
-108  
-109  
-112  
-113  
-121  
-123  
-124  
-128  
-131  
-138  
-141  
-148  
-150  
-151  
-158  
-161  
-168  
-200  
-220  
Command error [generic]  
Invalid character  
Syntax error [unrecognized command or data type]  
Invalid separator  
Data type error [e.g., “numeric or string expected, got block date”]  
GET not allowed  
Parameter not allowed [too many parameters]  
Missing parameter [too few parameters]  
Program mnemonic too long [maximum 12 characters]  
Undefined header [operation not allowed for this device]  
Invalid character in number [includes "9" in octal data, etc.]  
Numeric overflow [exponent too large; exponent magnitude >32 k]  
Too many digits [number too long; more than 255 digits received]  
Numeric data not allowed  
Invalid suffix [unrecognized units, or units not appropriate]  
Suffix not allowed  
Invalid character data [bad character, or unrecognized]  
Character data not allowed  
String data error  
Invalid string data [e.g., END received before close quote]  
String data not allowed  
Invalid block data [e.g., END received before length satisfied]  
Block data not allowed  
Execution error [generic]  
Parameter error  
Error Messages 65  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table 6-1. Summary of System Error Messages (continued)  
Error String [Description/Explanation/Examples]  
Error  
Number  
-222  
-223  
-241  
Data out of range [e.g., too large for this device]  
Too much data [out of memory; block, string, or expression too long]  
Hardware missing [device-specific]  
-310  
System error  
-330  
Self-test failed  
-350  
-400  
Too many errors [errors lost due to queue overflow]  
Query error [generic]  
-410  
-420  
-430  
-440  
Query INTERRUPTED [query followed by DAB or GET before response complete]  
Query UNTERMINATED [addressed to talk, incomplete programming message received]  
Query DEADLOCKED [too many queries in command string]  
Query UNTERMINATED [after indefinite response]  
66 Error Messages  
Download from Www.Somanuals.com. All Manuals Search And Download.  
A
SCPI Conformance Information  
Note  
See Chapter 3 - Language Dictionary for command syntax.  
SCPI Version  
This power module conforms to Version 1990.0.  
SCPI Confirmed Commands  
ABOR  
STAT:OPER:NTR?  
[SOUR]:VOLT[:LEV][:IMM][:AMPL)  
CAL:AUT  
CAL:STAT  
DISP[:WIND][:STAT]  
DISP[:WIND][:STAT]?  
INIT[:IMM]  
INIT:CONT  
INIT:CONT?  
MEAS:CURR[:DC]?  
MEAS:VOLT[:DC]?  
OUTP[:STAT]  
STAT:OPER:PTR  
STAT:OPER:PTR?  
STAT:PRES  
STAT:QUES[:EVEN]?  
STAT:QUES:COND?  
STAT:QUES:ENAB  
STAT:QUES:ENAB?  
[SOUR]:CURR[:LEV][:IMM][:AMPL]  
[SOUR]:CURR[:LEV][:IMM][:AMPL]?  
[SOUR]:CURR[:LEV]:TRIG[:AMPL]  
[SOUR]:CURR[:LEV]:TRIG[:AMPL]?  
[SOUR]:CURR:MODE  
[SOUR]:CURR:MODE?  
[SOUR]:CURR:PROT:STAT  
[SOUR]:CURR:PROT:STAT?  
[SOUR]:LIST:COUN  
[SOUR]:LIST:COUN?  
[SOUR]:LIST:CURR  
[SOUR]:LIST:CURR:POIN?  
[SOUR]:LIST:DWEL?  
[SOUR]:LIST:DWEL:POIN?  
[SOUR]:LIST:STEP  
[SOURI:VOLT[:LEV][:IMM][:AMPL]?  
[SOUR]:VOLT[:LEV][:TRIG][:AMPL]  
[SOUR]:VOLT[:LEV][:TRIG][:AMPL]?  
[SOUR]:VOLT:MODE  
[SOUR]:VOLT:MODE?  
[SOUR]:VOLT:PROT[:LEV]  
[SOUR]:VOLT:PROT[:LEV]?  
SYST:ERR?  
SYST:VERS?  
TRIG[:STAR][:IMM]  
TRIG:DEL  
TRIG:DEL?  
TRIG:LINK  
TRIG:LINK?  
TRIG:SOUR  
TRIG:SOUR?  
*CLS  
*ESE *ESE? *ESR?  
*IDN?  
*OPC *OPC? *OPT?  
*PSC *PSC?  
*RCL *RST  
*SAV *SRE *STB?  
*TRG *TST?  
OUTP[:STAT?]  
OUTP:PROT:CLE  
OUTP:PROT:DEL  
OUTP:PROT:DEL?  
OUTP:TTLT[:STAT]  
OUTP:TTLT[:STAT]?  
OUTP:TTLT:LINK  
OUTP:TTLT:LINK?  
OUTP:TTLT:SOUR  
OUTP:TTLT:SOUR?  
STAT:OPER[:EVEN]?  
STAT:OPER:COND?  
STAT:OPER:ENAB  
STAT:OPER:ENAB?  
STAT:OPER:NTR  
[SOUR]:LIST:STEP?  
[SOUR]:LIST:VOLT  
[SOUR]:LIST:VOLT:POIN?  
*WAI  
SCPI Approved Commands  
(None)  
SCPI Conformance Information 67  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Non-SCPI Commands  
CAL:CURR  
OUTP:DFI:SOUR?  
CAL:PASS  
CAL:SAVE  
CAL:VOLT  
OUTP:REL[:STAT]  
OUTP:REL[:STAT]?  
OUTP:REL:POL  
OUTP:DFI[:STAT]  
OUTP:DFI[:STAT]?  
OUTP:DFI:LINK  
OUTP:DFI:LINK?  
OUTP:DFI:SOUR  
OUTP:REL:POL?  
[SOUR]:LIST:STEP  
[SOUR]:LIST:STEP?  
[SOUR]:VOLT:SENS?  
68 SCPI Conformance Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
B
Application Programs  
This section contains seven example applications. For each application, there is:  
An overview of the application.  
Which MPS features are used to implement the application.  
The advantages and benefits of the MPS solution.  
The details of the implementation of the solution.  
A block diagram of the setup.  
A sample program listing in Agilent BASIC.  
A description of variations on the application.  
The following table lists what MPS features are used in each of the applications. It can be used as an index into this section.  
Application 1. Sequencing Multiple Modules During Power Up  
Application 2. Sequencing Multiple Modules to Power Down on Event  
Application 3. Controlling Output Voltage Ramp Up at Turn On  
Application 4. Providing Time-Varying Voltages  
Application 5. Providing Time-Varying Current Limiting  
Application 6. Output Sequencing Paced by the Computer  
Application 7. Output Sequencing Without Computer Intervention  
Application  
1
2
3
4
5
6
7
Lists  
l
20-point current List  
20-point voltage List  
Repetitive Lists  
Dwell time  
List Pacing  
Dwell-paced Lists  
Trigger-paced Lists  
Actions Due To A Change In Status  
Generate an SRQ  
l
l
l
l
l
l
l
l
l
l
l
l
l
l
l
l
l
l
Generate a trigger  
l
l
l
l
Disable the output  
Stop the List  
Triggers  
l
l
l
l
l
l
Change the voltage on trigger  
Trigger in/out from MPS backplane TTL  
Trigger  
Trigger on a GPIB trigger command  
Trigger delay  
Other Features  
Active downprogramming  
Overcurrent protection  
l
l
l
l
l
l
l
l
l
l
l
l
l
Application Programs 69  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Application 1. Sequencing Multiple Modules During Power Up  
Overview of Application  
When testing mixed signal devices, ± bias supply voltages are typically applied before logic bias supply voltages. For a  
device that is sensitive to when bias voltages are applied, the order of power up of multiple power modules can be  
controlled.  
For this example, the device requires three bias supplies, + 5 V for the logic circuits and ± 15 V for amplifier circuits. To  
properly power up the device, the supplies must be sequenced so that the ± 15 V are applied first and the + 5 V is applied 50  
ms later.  
The MPS can easily address this application through the use of triggers. The trigger will cause the modules to change from  
0 V, where they are not powering the DUT, to their final voltage. By delaying the response to the trigger, you can control  
when the module's output voltage changes. This means you can control the sequence of the modules during power up.  
MPS Features Used  
Change the voltage on trigger.  
Trigger in/out from MPS mainframe backplane TTL Trigger.  
Trigger on a GPIB trigger command Trigger delay.  
Trigger delay.  
Advantages/Benefits Of The MPS Solution  
By using trigger delay, the timing is accurate and repeatable.  
The sequence is simpler to program (no timing loops).  
The computer is not devoted to sequencing power modules.  
The computer does not provide timing for the sequence.  
One command initiates the sequence.  
Implementation Details  
How the MPS Implements The Sequence  
The computer sends a trigger command to the first module.  
The first module simultaneously sends a backplane trigger to other two modules and goes to + 15 V.  
The second module receives the backplane TTL Trigger and immediately goes to - 15 V.  
The third module receives the backplane TTL Trigger, delays 50 ms, and then goes to + 5 V.  
MPS Set Up  
Module in slot 0:  
The module is connected to + 15 V on the DUT.  
The initial voltage setting is 0 V.  
The module listens for the computer to send a trigger command.  
Upon receipt of the trigger command, the module goes to 15 V.  
Also upon receipt of the trigger command, the module generates a backplane TTL Trigger.  
Module in slot 1:  
The module is connected to - 15 V on the DUT.  
The initial voltage setting is 0 V.  
The module listens for a backplane TTL Trigger.  
Upon receipt of the trigger, the module goes to 15 V.  
70 Application Programs  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Module in slot 2:  
The module is connected to + 5 V on the DUT.  
The initial voltage setting is 0 V.  
The module listens for a backplane TTL Trigger.  
The trigger delay is programmed to 50 ms.  
Upon receipt of the trigger, the module waits the trigger delay time and then goes to 5 V.  
Variations On This Implementation  
1. The modules could be set to generate SRQ when the last module (+ 5 V) reaches its final output value. This would notify  
the computer that power has been applied to the DUT and the testing can begin.  
2. To provide a delay between the application of the + 15 V and the - 15 V bias, you can program different trigger delays  
into modules 2 and 3. The delay time will be relative to the module in slot 0.  
3. To get all three modules to apply power to the DUT at the same time, simply eliminate the trigger delay on the + 5 V  
module.  
4. When modules need to be connected in parallel to increase current, they will also need to be synchronized so that they all  
apply power simultaneously. To get modules in parallel to apply power at the same time, use the approach described in  
this example, but eliminate any trigger delays.  
Figure B1-1. Block Diagram of Application #1  
Application Programs 71  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Figure B1-2. Timing Diagram of Application #1  
72 Application Programs  
Download from Www.Somanuals.com. All Manuals Search And Download.  
10  
! APPLICATION #1: SEQUENCING MULTIPLE MODULES DURING POWER UP  
20  
30  
! PROGRAM: APP_1  
!
40  
ASSIGN @Slot0 TO 70500  
! SELECT CODE 7, MAINFRAME ADDRESS 05, SLOT 00  
50  
60  
ASSIGN @Slot1 TO 70501  
ASSIGN @Slot2 TO 70502  
! SELECT CODE 7, MAINFRAME ADDRESS 05, SLOT 01  
! SELECT CODE 7, MAINFRAME ADDRESS 05, SLOT 02  
70  
!
80  
90  
! SET UP MODULE IN SLOT 0 AS +15 V BIAS SUPPLY  
1
---------------------  
100  
110  
120  
130  
140  
OUTPUT@Slot0;"*RST;*CLS;STATUS:PRESET"  
OUTPUT@Slot0;”VOLT 0”  
OUTPUT@Slot0;”VOLT:TRIGGERED 15”  
OUTPUT@Stot0;”TRIGGER:SOURCE BUS”  
!
RESET AND CLEAR MODULE  
START AT 0 V  
GO TO 15 V ON TRIGGER  
TRIGGER SOURCE IS Agilent -18 'BUS'  
GENERATE BACKPLANE TTL TRIGGER WHEN GPIB 'BUS' TRIGGER IS  
RECEIVED  
!
!
!
OUTPUT@Stot0;”OUTPUT:TTLTRG:SOURCE BUS" !  
150  
160  
170  
180  
190  
200  
210  
220  
230  
240  
250  
260  
270  
280  
290  
300  
310  
320  
330  
340  
350  
360  
370  
380  
390  
400  
410  
420  
430  
440  
450  
460  
470  
480  
490  
500  
510  
520  
530  
540  
550  
OUTPUT@Slot0;”OUTPUT:TTLTRG:STATE ON"  
OUTPUT@Slot0;”OUTPUT ON"  
OUTPUT@Slot0;”INITIATE"  
!
! SET UP MODULE IN SLOT 1 AS -15 V BIAS SUPPLY  
1
OUTPUT@Slot1;"*RST;*CLS;STATUS:PRESET"  
OUTPUT@Slot1;"VOLT 0”  
OUTPUT@Slot1;"VOLT:TRIGGERED 15”  
OUTPUT@Slot1;"TRIGGER:SOURCE TTLTRG"  
OUTPUT @Slot1;"OUTPUT ON"  
OUTPUT @Slot1;"INITIATE"  
I
!
!
ENABLE BACKPLANE TTL TRIGGER DRIVE  
ENABLE OUTPUT  
ENABLE RESPONSE TO TRIGGER  
---------------------  
!
!
!
!
!
!
RESET AND CLEAR MODULE  
START AT 0 V  
GO TO 15 V ON TRIGGER  
TRIGGER SOURCE IS BACKPLANE TTL TRIGGER  
ENABLE OUTPUT  
ENABLE RESPONSE TO TRIGGER  
!
! SET UP MODULE IN SLOT 2 AS +5 V BIAS SUPPLY ---------------------  
!
OUTPUT @Slot2;"*RST;*CLS;STATUS:PRESET"  
OUTPUT @Slot2;"VOLT 0”  
OUTPUT @Slot2;"VOLT:TRIGGERED 5”  
OUTPUT @Slot2;"TRIGGER:SOURCE TTLTRG"  
OUTPUT @Slot2;"TRIGGER:DELAY 0.050”  
OUTPUT @Slot2;"OUTPUT ON"  
OUTPUT @Slot2;"INITIATE"  
!
!
!
!
!
!
!
RESET AND CLEAR MODULE  
START AT 0 V  
GO TO 5 V ON TRIGGER  
TRIGGER SOURCE IS BACKPLANE TTL TRIGGER  
50 ms TRIGGER DELAY  
ENABLE OUTPUT  
ENABLE RESPONSE TO TRIGGER  
!
! BEFORE TRIGGERING THE MODULES, DETERMINE IF THE MODULES ARE READY BY CHECKING FOR  
! 'WAITING FOR TRIGGER' (BIT 5 OF THE OPERATION STATUS REGISTER). IF THE LAST MODULE PROGRAMMED  
! IS READY THEN SO ARE THE OTHERS, SO JUST CHECK SLOT 2.  
!
!
!
!
YOU COULD ELIMINATE THIS STEP BY SIMPLY INSERTING A PAUSE IN THE PROGRAM. HOWEVER, BY  
CHECKING THE INSTRUMENT STATUS, YOU CAN AVOID TIMING PROBLEMS. ALSO, ANY OTHER OPERATIONS  
THAT TAKE TIME WILL GIVE THE MODULES A CHANCE TO COMPLETE PROCESSING.  
!
REPEAT  
OUTPUT @Slot2;"STATUS:OPERATION:CONDITION?"  
ENTER @Slot2;Condition_data  
UNTIL BIT(Condition_data,5)  
!
! TRIGGER MODULE IN SLOT 0 TO BEGIN SEQUENCING THE 3 MODULES TO POWER UP  
!
OUTPUT @Slot0;"*TRG”  
!
TEST FOR BIT 5 = TRUE  
!
SEND Agilent -1B 'BUS' TRIGGER  
!
END  
Figure B1-3. Agilent BASIC Program Listing for Application #1  
Application Programs 73  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Application 2. Sequencing Multiple Modules to Power Down on Event  
Overview Of Application  
When testing devices, such as some GaAs and ECL devices that are sensitive to when bias voltages are removed, the order  
of power-down of multiple power modules can be controlled. The power-down sequence can be initiated by an event, such  
as a change in power module status, fault condition, detection of a TTL signal, etc.  
For this example, there are three supplies + 5 V and ± 15 V. (See previous application for how to generate a power up  
sequence.) Once the power has been applied to the DUT, the modules can be reprogrammed to perform the power down  
sequence. The power down sequence is initiated when a fault in the DUT draws excessive current from the power module,  
causing the module to change from CV to CC. To prevent damage to the DUT, it is necessary to remove the + 5 V first,  
then the ± 15 V modules 15 ms later.  
Once again, MPS triggering can solve the application. In this scenario, the CV-to-CC crossover event will be used as the  
trigger source. The trigger will cause the modules, in the correct order, to change from their programmed voltages down to  
0 V.  
MPS Features Used  
Generate a trigger on a change in internal status.  
Change the voltage on trigger.  
Trigger in/out from MPS mainframe backplane TTL Trigger o Trigger delay.  
Active downprogramming.  
Advantages/Benefits Of The MPS Solution  
By using the modules' change in status to automatically generate a trigger, the computer is not devoted to polling the  
modules to detect a change in state.  
By letting each module monitor its status, the CC condition will generate a response faster than if the computer was polling  
the module to detect a change in state.  
The sequence is simpler to program (no timing loops).  
By using trigger delay, the timing is accurate and repeatable because the computer does not provide timing for the sequence.  
The active downprogrammers in the module output can quickly discharge the module's output capacitors and any  
capacitance in the DUT.  
Implementation Details  
How The MPS Implements The Solution  
All modules are set to listen for a backplane TTL Trigger.  
When any module detects a change in status from CV to CC, it sends out a backplane TTL Trigger.  
When the + 5 V module receives the trigger, it immediately goes to 0 V.  
When the + 15 V and - 15 V modules receive the trigger, they wait the trigger delay time and then go to 0 V.  
Note  
Any module can generate both the backplane TTL Trigger signal and be triggered by that same signal.  
74 Application Programs  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MPS Set Up  
Module in slot 0:  
The module is connected to + 15 V on the DUT.  
The initial voltage setting is 15 V.  
The module monitors its status.  
The module will generate a backplane TTL Trigger on CV-to-CC crossover.  
The module listens for a backplane TTL Trigger.  
The trigger delay is programmed to 15 ms.  
Upon receipt of the trigger, the module waits the trigger delay time and then goes to 0 V.  
Module in slot 1:  
The module is connected to supply - 15 V to the DUT.  
The initial voltage setting is 15 V  
The module monitors its status.  
The module will generate a backplane TTL Trigger on CV-to-CC crossover.  
The module listens for backplane TTL Trigger.  
The trigger delay is programmed to 15 ms.  
Upon receipt of the trigger, the module waits the trigger delay time and then goes to 0 V.  
Module in slot 2.  
The module is connected to supply + 5 V to the DUT.  
The initial voltage setting is 5 V.  
The module monitors its status.  
The module will generate a backplane TTL Trigger on CV-to-CC crossover.  
The module listens for backplane TTL Trigger.  
Upon receipt of the trigger, the module immediately goes to 0 V.  
Variations On This Implementation  
1. The modules could be set to generate SRQ when the last module reaches 0 V. This could notify the computer that power  
has been removed from the DUT.  
2. The modules could be set to generate a DFI (Discrete Fault Indicator) signal on the MPS rear panel on a change in status.  
This signal could be used to shut down other power modules, to flash an alarm light, or to sound a buzzer. This could  
also be routed to other instruments to signal them to stop making measurements.  
3. To get all three modules to remove power from the DUT at the same time, simply eliminate the trigger delay on the ± 15  
V modules.  
4. To provide a delay between the removal of the three bias voltages, you can program a different trigger delay into each  
module.  
Application Programs 75  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Figure B2-1. Block Diagram of Application #2  
Figure B2-2. Timing Diagram of Application #2  
76 Application Programs  
Download from Www.Somanuals.com. All Manuals Search And Download.  
10  
! APPLICATION #2: SEQUENCING MULTIPLE MODULES TO POWER DOWN ON EVENT  
20  
30  
! PROGRAM: APP_2  
!
40  
50  
60  
ASSIGN @Slot0 TO 70500  
ASSIGN @Slot1 To 70501  
ASSIGN @Slot2 TO 70502  
!
!
!
SELECT CODE 7, MAINFRAME ADDRESS 05, SLOT 00  
SELECT CODE 7, MAINFRAME ADDRESS 05, SLOT 01  
SELECT CODE 7, MAINFRAME ADDRESS 05, SLOT 02  
70  
!
80  
90  
! SET UP MODULE IN SLOT 0 AS +15 V BIAS SUPPLY  
!
---------------------  
100  
110  
120  
130  
140  
150  
160  
170  
180  
190  
200  
210  
220  
230  
240  
250  
260  
270  
280  
290  
300  
310  
320  
330  
340  
350  
360  
370  
380  
390  
400  
410  
420  
430  
440  
450  
460  
470  
480  
490  
500  
510  
520  
OUTPUT @Slot0;"*RST;*CLS;STATUS:PRESET"  
OUTPUT @Slot0;”CURR .5”  
OUTPUT @Slot0;”VOLT 15”  
OUTPUT @Slot0;”VOLT:TRIGGERED 0”  
OUTPUT @Slot0;”TRIGGER:SOURCE TTLTRG"  
OUTPUT @Slot0;”TRIGGER:DELAY .015”  
OUTPUT @Slot0;”INITIATE"  
!
RESET AND CLEAR MODULE  
!
!
!
!
!
START AT 15 V  
GO TO 0 V ON TRIGGER  
TRIGGER SOURCE IS TTL TRIGGER  
15 ms TRIGGER DELAY  
ENABLE RESPONSE TO TRIGGER  
OUTPUT @Slot0;"OUTPUT:TTLTRG:SOURCE LINK" ! GENERATE A BACKPLANE TTL TRIGGER  
OUTPUT @Slot0;”OUTPUT:TTLTRG:LINK 'CC’ “  
OUTPUT @Slot0;”OUTPUT:TTLTRG:STATE ON"  
OUTPUT @Slot0;"OUTPUT ON"  
!
! SET UP MODULE IN SLOT 1 AS -15 V BIAS SUPPLY ---------------------  
!
OUTPUT @Slot1;"*RST;*CLS;STATUS:PRESET"  
OUTPUT @Slot1;"CURR .5”  
OUTPUT @Slot1;"VOLT 15”  
OUTPUT @Slot1;"VOLT:TRIGGERED 0”  
OUTPUT @Slot1;"TRIGGER:SOURCE TTLTRG"  
OUTPUT @Slot1;"TRIGGER:DELAY .015”  
OUTPUT @Slot1;"INITIATE"  
!
!
!
WHEN A CV-TO-CC TRANSITION OCCURS  
ENABLE TTL TRIGGER DRIVE  
ENABLE OUTPUT  
!
RESET AND CLEAR MODULE  
!
!
!
!
!
START AT 15 V  
GO TO 0 V ON TRIGGER  
TRIGGER SOURCE IS BACKPLANE TTL TRIGGER  
15 ms TRIGGER DELAY  
ENABLE RESPONSE TO TRIGGER  
OUTPUT @Slot1;"OUTPUT:TTLTRG:SOURCE LINK" ! GENERATE A BACKPLANE TTL TRIGGER  
OUTPUT @Slot1;"OUTPUT:TTLTRG:LINK 'CC’ “  
OUTPUT @Slot1;"OUTPUT:TTLTRG:STATE ON"  
OUTPUT @Slot1;"OUTPUT ON"  
!
! SET UP MODULE IN SLOT 2 AS +5 V BIAS SUPPLY  
!
OUTPUT @Slot2;"*RST;*CLS;STATUS:PRESET"  
OUTPUT @Slot2;"CURR .5”  
OUTPUT @Slot2;"VOLT 5”  
OUTPUT @Slot2;"VOLT:TRIGGERED 0”  
OUTPUT @Slot2;"TRIGGER:SOURCE TTLTRG"  
OUTPUT @Slot2;"INITIATE"  
OUTPUT @Slot2;"OUTPUT:TTLTRG:SOURCE LINK"  
OUTPUT @Slot2;"OUTPUT:TTLTRG:LINK 'CC’ “  
OUTPUT @Slot2;"OUTPUT:TTLTRG:STATE ON"  
OUTPUT @Slot2;"OUTPUT ON"  
!
!
!
!
WHEN A CV-TO-CC TRANSITION OCCURS  
ENABLE TTL TRIGGER DRIVE  
ENABLE OUTPUT  
----------------------  
!
RESET AND CLEAR MODULE  
!
!
!
!
!
!
!
!
START AT 5 V  
GO TO 0 V ON TRIGGER  
TRIGGER SOURCE IS BACKPLANE TTL TRIGGER  
ENABLE RESPONSE TO TTL TRIGGER  
GENERATE A BACKPLANE TTL TRIGGER  
WHEN A CV-TO-CC TRANSITION OCCURS  
ENABLE TTL TRIGGER DRIVE  
ENABLE OUTPUT  
! THE POWER MODULES ARE NOW SET UP TO IMPLEMENT THE POWER DOWN ON EVENT.  
! ANY TIME ANY MODULE GOES INTO CC, THE SEQUENCE WILL OCCUR.  
!
END  
Figure B2-3. Agilent BASIC Program Listing for Application #2  
Application Programs 77  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Application 3. Controlling Output Voltage Ramp Up at Turn On  
Overview Of Application  
When control over the rate of voltage ramp up at turn-on of the power module output is required, the desired shape can be  
approximated by downloading and executing a series of voltage and dwell time points.  
For this example, you need to program the power module to change its output from 2 volts to 10 volts, slewing through the 8  
volt transition in 0.5 seconds. This results in a turn-on ramp-up of 16 V per second.  
The MPS can create this voltage versus time characteristic using Lists. The desired characteristic (in this case, linear) is  
simulated using the 20 available voltage points. To determine the value of each point in the transition, simply divide the  
change in voltage by 20. To determine the dwell time of each voltage point, divide the total transition time by 19. After the  
List has been executed, the module will continue to output the final value (in this case, 10 volts) until the output has been  
reprogrammed to another value. Note that the dwell-time of the last point is not part of the transition time.  
To determine the slowest ramp up (longest transition time) that can be generated, you must consider how smooth you need  
the voltage versus time characteristic to be. As the dwell time associated with each point gets longer, the output voltage will  
become more like a "stair step" and less like a linear transition. (see Figure B3-1)  
To determine the fastest ramp up (shortest transition time) that can be generated, you must consider the minimum dwell time  
specification (10 ms) and the maximum risetime of specification the power module (20 ms). If you program 10 ms dwell  
times, the power module will not be able to reach its output voltage before the next voltage point is output. (see Figure B3-  
2)  
MPS Features Used  
20-point voltage List.  
Dwell time.  
Dwell-paced Lists.  
Advantages/Benefits Of The MPS Solution  
By using Lists, the module changes its output voltage automatically, so that the computer is not devoted to reprogramming  
the output voltage.  
The outputs can change faster when dwell paced than when the computer must explicitly reprogram each change.  
The sequence is simpler to program (no timing loops). By using dwell times, the timing of each point is accurate and  
repeatable.  
The computer does not provide timing for the sequence. For negative-going ramps, the active downprogrammers in the  
module output can quickly discharge the module’s output capacitors and any capacitance in the DUT when negative going  
ramps are required.  
Implementation Details  
How the MPS Implements The Sequence  
The module is programmed to List mode.  
The module will execute a dwell-paced List.  
The 20 voltage points are downloaded to the module.  
The 20 dwell times are downloaded to the module.  
When the transition must occur, the module is triggered by the computer.  
The module output ramps under its own control.  
78 Application Programs  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Figure B3-1. Simulating a Slow Voltage Ramp  
Figure B3-2. Simulating a Fast Voltage Ramp  
Variations On This Implementation  
1. The module could be set to begin ramping in response to an external or backplane TTL Trigger.  
2. The module could be set to generate SRQ when it has finished its transition. This would notify the computer that the  
voltage is at the proper level.  
Application Programs 79  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. The module could be set to generate an external trigger when it has finished its transition. This trigger could be routed to  
other instruments as a signal to start making measurements.  
4. Multiple modules could be programmed to slew together in response to the computer trigger command.  
5. The module could be set to generate an external trigger for each point in the transition. This trigger could be routed to  
other instruments as a signal to take a measurement at various supply voltages. (see application #7)  
6. Many voltage versus time characteristics can be generated by varying the voltage values and the dwell times in the List.  
Figure 3-3. Generating the Desired Voltage Ramp for Application #3  
80 Application Programs  
Download from Www.Somanuals.com. All Manuals Search And Download.  
10  
! APPLICATION #3: CONTROLLING VOLTAGE RAMP UP AT TURN ON  
20  
30  
! PROGRAM: APP_3  
!
40  
50  
ASSIGN @Slot0 To 70500  
!
!
SELECT CODE 7, MAINFRAME ADDRESS 05, SLOT 00  
60  
70  
80  
OPTION BASE 1  
DIM V_Step(20)  
Vstart=2  
!
!
!
!
!
ARRAY TO HOLD THE VOLTAGE RAMP STEPS  
START VOLTAGE FOR RAMP  
STOP VOLTAGE FOR RAMP  
SECONDS TO CHANGE FROM Vstart TO Vstop  
IN SECONDS  
90  
Vstop=10  
100  
110  
120  
130  
140  
150  
160  
170  
180  
190  
200  
210  
220  
230  
240  
250  
260  
270  
280  
290  
300  
310  
320  
330  
340  
350  
360  
370  
380  
390  
400  
410  
420  
430  
440  
450  
460  
470  
Ramp_time=.5  
Dwell=Ramp_time/19  
!
! SINCE THE OUTPUT STAYS AT THE LAST VOLTAGE POINT AFTER ITS DWELL TIME EXPIRES, THE DWELL TIME OF THE  
! LAST POINT IS NOT PART OF THE TRANSITION TIME. THEREFORE, DIVIDE THE TOTAL TIME BY 19 POINTS, NOT 20.  
! ALSO, YOU ONLY NEED TO DOWNLOAD 1 DWELL TIME. IF THE MODULERECEIVES ONLY 1 DWELL TIME, IT ASSUMES  
! YOU WANT THE SAME DWELL TIME FOR EVERY POINT IN THE LIST.  
!
FOR I=1 TO 20  
V_step(l)=Vstart+(((Vstop-Vstart)/20)*I)  
NEXT I  
!
CALCULATES VOLTAGE LIST POINTS  
!
OUTPUT @Slot0;"*RST;*CLS;STATUS:PRESET"  
OUTPUT @Slot0;”VOLT ";Vstart  
OUTPUT @Slot0;”CURR .1”  
OUTPUT @Slot0;"OUTPUT ON"  
OUTPUT @Slot0;”VOLT:MODE LIST"  
OUTPUT @Slot0;”LIST:VOLT ";V_step(*)  
OUTPUT @Slot0;”LIST:DWELL ";Dwell  
OUTPUT @Slot0;”LIST:STEP AUTO"  
OUTPUT @Slot0;”INITIATE"  
!
!
!
RESET AND CLEAR MODULE  
START RAMP AT Vstart  
!
!
!
!
!
!
ENABLE OUTPUT  
SET TO GET VOLTAGE FROM LIST  
DOWNLOAD VOLTAGE POINTS  
DOWNLOAD 1 DWELL TIME  
DWELL-PACED LIST  
ENABLE TRIGGER TO START LIST  
! BEFORE TRIGGERING THE MODULE, DETERMINE IF IT IS READY BY CHECKING FOR  
! 'WAITING FOR TRIGGER' (BIT 5 OF THE OPERATION STATUS REGISTER).  
!
! YOU COULD ELIMINATE THIS STEP BY SIMPLY INSERTING A PAUSE IN THE PROGRAM. HOWEVER, BY  
! CHECKING THE INSTRUMENT STATUS, YOU CAN AVOID TIMING PROBLEMS. ALSO, ANY OTHER OPERATIONS  
! THAT TAKE TIME WILL GIVE THE MODULE A CHANCE TO COMPLETE PROCESSING.  
!
REPEAT  
OUTPUT @Slot0;”STATUS:OPERATION:CONDITION?"  
ENTER @Slot0;Condition_data  
UNTIL BIT(Condition_data,5)  
!
!
TEST FOR BIT 5 = TRUE  
! SEND TRIGGER COMMAND TO START LIST AND GENERATE THE VOLTAGE RAMP  
!
OUTPUT @Slot0;”TRIGGER:IMMEDIATE"  
!
!
THIS IS AN IMMEDIATE TRIGGER, WHICH IS ALWAYS ACTIVE.  
THEREFORE, IT DOES NOT NEED TO BE SELECTED AS A TRIGGER  
SOURCE.  
480  
490  
!
END  
Figure B3-4. Agilent BASIC Program Listing for Application #3  
Application Programs 81  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Application 4. Providing Time-Varying Voltages  
Overview of Application  
To burn-in devices using thermal or mechanical cycling/stress, cyclical time-varying voltage is provided by programming a  
set of voltage and dwell time points that repetitively sequence over time.  
For this example, the power module must provide the repetitive waveform shown in Figure B4-1. This time-varying voltage  
will be applied to a hybrid IC. By continually cycling the voltage from 0 to 7 volts over a 33 second interval, the hybrid is  
given time to heat up and undergo thermal and mechanical stress as the welds inside the hybrid expand and contract, and  
then subsequently cool down.  
Figure B4-1. Voltage Waveform for Application #4  
In addition to generating the cyclical voltage, it is desirable to have the power module notify the computer should the device  
fail and stop the cycling. Since the module is monitoring test status, the computer is free to perform other tests.  
The MPS can address this application using dwell-paced repetitive Lists. This application could be thought of as a simple  
power arbitrary waveform generator. To get the desired time-varying voltage, you must be able to describe the waveform in  
20 discrete voltage points, with each point ranging from 10 ms to 65 seconds. This range of dwell times determines the  
range of frequencies (or time rate of change) of the voltage waveform to be generated.  
Once the waveform has been described, it is downloaded to the module. Upon being triggered, it will repetitively generate  
the waveform without computer intervention.  
The module will also be set up to generate an SRQ and stop the voltage cycling of the hybrid should fail. If the hybrid fails  
by shorting, the module will go into CC. This change in status will cause the module to protect the DUT by disabling the  
output, which will stop the test and generate an SRQ. (Open circuit failures will not be detected. Since failures of this type  
are less likely to have destructive consequences, detection is not required.)  
82 Application Programs  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MPS Features Used  
20-point voltage List.  
Repetitive Lists.  
Dwell time.  
Dwell-paced Lists.  
Generate an SRQ on a change in internal status.  
Disable the output on a change in internal status.  
Stop the List on a change in internal status.  
Trigger on a GPIB trigger command.  
Overcurrent protection.  
Active downprogramming.  
Advantages/Benefits Of The MPS Solution  
By using Lists, the module changes its output voltage automatically, so that the computer is not devoted to reprogramming  
the output voltage.  
The output can change faster when dwell paced than when the computer must explicitly reprogram each change.  
Overcurrent protection can disable the output before the DUT is damaged.  
By letting each module monitor its status, the CC condition will be responded to faster than if the computer was responsible  
for stopping the test.  
The sequence is simpler to program (no timing loops),  
By using dwell times, the timing of each point is accurate and repeatable because the computer does not provide timing for  
the sequence.  
When the output is disabled, the active downprogrammers in the module output can quickly discharge the module’s output  
capacitors and any capacitance in the DUT.  
Implementation Details  
How The MPS Implements The Sequence  
The module is programmed to List mode.  
The module will execute a dwell-paced List.  
The 3 voltage points are downloaded to the module.  
The 3 dwell times are downloaded to the module.  
To begin the cycling, the module is triggered by the computer.  
The module continuously generates the voltage waveform.  
The module continuously monitors its status.  
If the module goes into CC, the overcurrent protection disables the output.  
The module generates an SRQ when the overcurrent protection occurs.  
Module set up  
Set voltage mode to List.  
Download voltage List.  
Download dwell times.  
Set Lists to dwell paced.  
Set Lists to infinitely repeat.  
Enable status monitoring of overcurrent condition.  
Enable overcurrent protection.  
Enable SRQ generation on overcurrent protection occurrence.  
Application Programs 83  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Variations On This Implementation  
1. The module could be set to begin generating the waveform in response to an external or backplane TTL Trigger.  
2. The module could be set to generate external triggers for each point in the List. This trigger could be routed to other  
instruments to synchronize external measurements to the change in voltage. (see application #7) Using this technique,  
parametric measurements could be made on the device during the thermal cycling.  
3. Multiple modules could be programmed to cycle together in response to the computer trigger command.  
4. To determine how many times the hybrid was cycled before it failed, you can use the SRQ (that was generated when the  
hybrid failed and the module went into CC) to timestamp the failure. The elapsed time will give the number of cycles  
executed.  
84 Application Programs  
Download from Www.Somanuals.com. All Manuals Search And Download.  
10  
! APPLICATION #4: PROVIDING TIME-VARYING VOLTAGES  
20  
! PROGRAM: APP_4  
30  
!
40  
ASSIGN Slot0 TO 70500  
!
SELECT CODE 7, MAINFRAME ADDRESS 05, SLOT 00  
50  
!
60  
! INITIALIZE THE MODULE  
70  
!
80  
90  
OUTPUT @Slot0;"*RST;*CLS;STATUS:PRESET"  
OUTPUT @Slot0;”VOLT 0”  
OUTPUT @Slot0;”CURR .1”  
OUTPUT @Slot0;”OUTPUT ON"  
!
! SET UP OVERCURRENT PROTECTION (OCP) AND GENERATE SRQ ON OCP TRIP  
!
OUTPUT @Slot0;”CURRENT:PROTECTION:STATE ON"  
OUTPUT @Slot0;”OUTPUT:PROTECTION:DELAY 0”  
OUTPUT @Slot0;”STATUS:QUESTIONABLE:ENABLE 2”  
!
!
!
!
RESET AND CLEAR MODULE  
START TEST AT 0 V  
SET CURRENT LIMIT  
ENABLE OUTPUT  
100  
110  
120  
130  
140  
150  
160  
170  
180  
190  
!
!
!
!
!
ENABLE OCP  
NO DELAY BEFORE PROTECTION OCCURS  
ENABLE DETECTION OF OC CONDITION IN THE  
QUESTIONABLE REGISTER, WHERE OC = BIT 1 = VALUE 2.  
ENABLES DETECTION ON POSITIVE TRANSITION, I.E.,  
GOING INTO OC.  
OUTPUT @Slot0;”STATUS:QUESTIONABLE:PTRANSITION 2”  
OUTPUT @Slot0;"*SRE 8”  
200  
210  
220  
230  
240  
250  
260  
270  
280  
290  
300  
310  
320  
330  
340  
350  
360  
370  
380  
390  
400  
410  
420  
430  
440  
450  
460  
470  
480  
490  
!
!
!
ENABLES THE SERVICE REQUEST REGISTER TO GENERATE  
AN SRQ WHEN ANY EVENT IN THE QUESTIONABLE REGISTER  
IS ASSERTED. THE QUESTIONABLE REGISTER = BIT 3= VALUE 8.  
!
! SET UP THE VOLTAGE LIST  
!
OUTPUT @Slot0;”VOLT:MODE LIST"  
OUTPUT @Slot0;”LIST:VOLT 5,7,0”  
OUTPUT @Slot0;”LIST:DWELL 1,2,30”  
OUTPUT @Slot0;’LIST:STEP AUTO"  
OUTPUT @Slot0;”LIST:COUNT INF"  
OUTPUT @Slot0;”INITIATE"  
!
!
!
!
!
!
!
SET TO GET VOLTAGE FROM LIST  
DOWNLOAD VOLTAGE POINTS  
DOWNLOAD DWELL TIMES  
DWELL-PACED LIST  
CONTINUOUSLY REPEAT LIST (INF = INFINITE)  
ENABLE TRIGGER TO START LIST  
!
! BEFORE TRIGGERING THE MODULE, DETERMINE IF IT IS READY BY CHECKING FOR  
! 'WAITING FOR TRIGGER' (BIT 5 OF THE OPERATION STATUS REGISTER).  
! YOU COULD ELIMINATE THIS STEP BY SIMPLY INSERTING A PAUSE IN THE PROGRAM. HOWEVER, BY  
! CHECKING THE INSTRUMENT STATUS, YOU CAN AVOID TIMING PROBLEMS. ALSO, ANY OTHER OPERATIONS  
! THAT TAKE TIME WILL GIVE TNE MODULE A CHANCE TO COMPLETE PROCESSING.  
!
REPEAT  
OUTPUT @Slot0;”STATUS:OPERATION:CONDITION?"  
ENTER @Slot0;Condition_data  
UNTIL BIT(Condition_data,5)  
!
TEST FOR BIT 5 = TRUE  
!
! SEND Agilent -1B TRIGGER COMMAND TO START LIST  
!
OUPUT @Slot0;”TRIGGER:IMMEDIATE"  
!
!
THIS IS AN IMMEDIATE TRIGGER, WHICH IS ALWAYS ACTIVE.  
THEREFORE, IT DOES NOT NEED TO BE SELECTED AS A TRIGGER  
SOURCE.  
500  
510  
!
END  
Figure B4-2. Agilent BASIC Programming Listing for Application #4  
Application Programs 85  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Application 5. Providing Time-Varying Current Limiting  
Overview Of Application  
To provide current limit protection which varies as a function of time, multiple thresholds on current limit are required.  
Having multiple thresholds can provide a high limit to protect the DUT during its power-up in-rush with automatic  
switchover to a lower limit to protect the DUT during its steady state operation.  
For this example, the DUT is a printed circuit assembly. This assembly is being tested prior to installation in the end  
product. The module provides power to the assembly, which will undergo a functional test. The assembly has capacitors  
on-board, and when power is applied, the in-rush current approaches 4 A. After the capacitors charge, which takes about  
500 milliseconds, the steady state current settles to 600 mA. See Figure B5-1.  
The MPS can address this application using dwell-paced Lists. In this case, the List will consist of a set of current limits  
and dwell times, because the voltage will remain constant throughout the test.  
Once power has been applied, the first current limit, which provides protection to a shorted DUT while still allowing high  
current in-rush to occur, will remain in effect for the dwell time. Then the current limit will switch to its next setting in the  
List. The result is a current limit which changes with time and provides protection as the DUT current requirements drop  
off to their steady state value. When the dwell time expires for the last current limit in the List, the current limit stays at this  
value until reprogrammed. Thus, the actual value of the last dwell time is not important. The last current List point would  
be the current limit for the steady state operation during the test of the DUT. See Figure B5-2 for how the MPS implements  
this protection.  
Throughout List execution, overcurrent protection will be enabled. If at any time the module goes into CC, the output will  
be disabled, the test stopped, and the DUT protected.  
MPS Features Used  
20-point current List.  
Dwell time.  
Dwell-paced Lists.  
Disable the output on a change in internal status.  
Stop the List on a change in internal status.  
Change the voltage on trigger.  
Trigger on a GPIB trigger command.  
Overcurrent protection.  
Active downprogramming.  
Advantages/Benefits Of The MPS Solution  
By using Lists, the module changes its current limit automatically, so that the computer is not devoted to reprogramming the  
current limit.  
The output can change faster when dwell paced than when the computer must explicitly reprogram each change.  
Overcurrent protection can disable the output before the DUT is damaged.  
By letting the modules monitor status, the CC condition will be responded to faster than if the computer was responsible for  
stopping the test.  
The sequence is simpler to program (no timing loops).  
By using dwell times, the timing of each point is accurate and repeatable because the computer does not provide timing for  
the sequence.  
When the output is disabled, the active downprogrammers in the module output can quickly discharge the module’s output  
capacitors and any capacitance in the DUT.  
86 Application Programs  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Figure B5-1. Typical DUT Current vs. Time  
Figure B5-2. Desired Current vs. Time  
Application Programs 87  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Implementation Details  
How The MPS Implements The Sequence.  
The module is programmed to current List mode.  
The module will execute a dwell-paced current List.  
The current limit List points are downloaded to the module.  
The dwell times are downloaded to the module.  
To begin powering the DUT, the module is triggered by the computer. This one trigger causes the current List to begin  
executing and the voltage to go to its programmed value.  
The module steps through the current limit List.  
The module continuously monitors its status.  
If the modules goes into CC, the overcurrent protection disables the output.  
Module Set Up  
Set the current mode to List.  
Download current List.  
Download the dwell times.  
Set the List to be dwell paced.  
Enable overcurrent protection.  
The initial voltage setting is 0 V.  
The module listens for the computer to send a trigger command.  
Upon receipt of the trigger command, the module goes to 12 V.  
Also upon receipt of the trigger command, the module begins executing its current limit List.  
Variations On This Implementation  
1. The module could be set to begin applying power in response to an external or backplane TTL Trigger.  
2. Multiple modules could be programmed to cycle together in response to the computer trigger command. Each module  
could have unique current limits, voltage settings, and dwell times.  
3. The module could be set to generate an SRQ if the overcurrent protection disables the output.  
4. The module could be set to generate an external or backplane TTL Trigger if the overcurrent protection disables the  
output.  
5. The in-rush current can be controlled using the current limit settings of the current List. Instead of setting the current  
limit slightly above 4 A, it could be set at a much lower value. This would limit the in-rush current to the value in the  
List. It would take longer to charge the capacitors on the assembly, but the inrush condition would be controlled. In this  
variation, the overcurrent protection could not be used, because you want the module to be in CC.  
88 Application Programs  
Download from Www.Somanuals.com. All Manuals Search And Download.  
10  
! APPLICATION #5: PROVIDING TIME-VARYING CURRENT LIMITING  
20  
! PROGRAM: APP_5  
30  
!
40  
50  
DIM C_limit$[50],Dwell[50]  
!
60  
70  
80  
C_limit$=”4.1, 3.0, 2.0 , 1.0, 0.7”  
Dwell$="0.2, 0.05, 0.1, 0.15, 0.1”  
!
!
!
CURRENT LIMIT DATA  
DWELL TIME DATA  
90  
ASSIGN @Slot0 To 70500  
!
!
SELECT CODE 7, MAINFRAME ADDRESS 05, SLOT 00  
100  
110  
120  
130  
140  
150  
160  
170  
180  
190  
200  
210  
220  
230  
240  
250  
260  
270  
280  
290  
300  
310  
320  
330  
340  
350  
360  
370  
380  
390  
OUTPUT @Slot0;"*RST;*CLS;STATUS:PRESET"  
OUTPUT @Slot0;”VOLT 0”  
OUTPUT @Slot0;”OUTPUT ON"  
OUTPUT @Slot0;”CURRENT:PROTECTION:STATE ON"  
OUTPUT @Slot0;"OUTPUT:PROTECTION:DELAY 0” !  
OUTPUT @Slot0;”CURRENT:MODE LIST"  
OUTPUT @Slot0;”LIST:CURRENT ";C _limit$  
OUTPUT @Slot0;”LIST:DWELL ";Dwell$  
OUTPUT @Slot0;”LIST:STEP AUTO"  
OUTPUT @Slot0;”VOLT:TRIGGERED 12”  
OUTPUT @Slot0;”INITIATE"  
!
!
!
RESET AND CLEAR MODULE  
START TEST AT 0 V  
ENABLE OUTPUT  
!
ENABLE OCP  
NO DELAY BEFORE PROTECTION OCCURS  
SET TO GET CURRENT FROM LIST  
DOWNLOAD CURRENT POINTS  
DOWNLOAD DWELL TIMES  
DWELL-PACED LIST  
GO TO 12 V WHEN TRIGGERED  
ENABLE TRIGGER TO START LIST AND APPLY 12 V  
!
!
!
!
!
!
!
! BEFORE TRIGGERING THE MODULE, DETERMINE IF IT IS READY BY CHECKING FOR  
! 'WAITING FOR TRIGGER' (BIT 5 OF THE OPERATION STATUS REGISTER).  
!
! YOU COULD ELIMINATE THIS STEP BY SIMPLY INSERTING A PAUSE IN THE PROGRAM. HOWEVER, BY  
! CHECKING THE INSTRUMENT STATUS, YOU CAN AVOID TIMING PROBLEMS. ALSO, ANY OTHER OPERATIONS  
! THAT TAKE TIME WILL GIVE THE MODULE A CHANCE TO COMPLETE PROCESSING.  
!
REPEAT  
OUTPUT @Slot0;”STATUS:OPERATION:CONDITION?"  
ENTER @Slot0;Condition_data  
UNTIL BIT(Condition_data,5)  
!
! SEND Agilent -1BTRIGGER COMMAND TO START LIST AND APPLY 12 V  
!
OUTPUT @Slot0;”TRIGGER:IMMEDIATE"  
!
TEST FOR BIT 5 = TRUE  
!
!
!
THIS IS AN IMMEDIATE TRIGGER, WHICH IS  
ALWAYS ACTIVE. THEREFORE, IT DOES NOT NEED  
TO BE SELECTED AS A TRIGGER SOURCE.  
END  
Figure B5-3. Agilent BASIC Program Listing for Application #5  
Application Programs 89  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Application 6. Output Sequencing Paced by the Computer  
Overview Of Application  
When performing bias supply margin testing, throughput can be maximized by eliminating the command processing time  
associated with reprogramming all outputs for each set of limit conditions. Instead, multiple sets of bias limit conditions can  
be downloaded to the power modules during test system initialization. During the testing, the computer can use a single  
command to simultaneously signal all power modules to step through each test condition.  
In this example, the DUT requires + 5 V and ± 12 V. The DUT is tested to ensure proper operation at marginal supply  
voltages. The margin specified is ± 5 % of nominal voltage. At each of the combinations given below, the computer first  
sets up the three modules and makes a measurement on the DUT. The combinations to be tested are:  
Nominal 5 V  
4.75 V  
5 V  
Nominal + 12 V  
12 V  
Nominal - 12 V  
-12 V  
12 V  
-12 V  
5.25 V  
5 V  
5 V  
5 V  
5 V  
12 V  
-12 V  
-12 V  
-12 V  
- 11.4 V  
- 12.6 V  
11.4 V  
12.6 V  
12 V  
12 V  
When conducting this test, the modules will need to be reprogrammed 21 times and seven measurements made. The  
command processing time could slow down this test.  
The MPS can be used to increase throughput. By downloading all of the combinations into the three modules, each setting  
can be quickly stepped through by triggering all modules to change to their next voltage setting and then taking a  
measurement from the DUT. This permits testing without command processing overhead.  
MPS Features Used  
20-point voltage List.  
Trigger-paced Lists.  
Trigger in/out from MPS mainframe backplane TTL Trigger.  
Trigger on a GPIB trigger command.  
Advantages/Benefits Of The MPS Solution  
By using Lists, the module changes its voltage without delays due to processing the command to change the output voltage.  
By using triggers, all three outputs can be changed with one command.  
The computer loop to change the settings and take a measurement is simplified, because you do not have to explicitly  
reprogram each module output. Instead, the loop becomes "Trigger" and "Measure".  
Implementation Details  
How The MPS Implements The Sequence  
The following steps are performed for each point in the List:  
The computer sends a trigger command to the first module.  
The first module simultaneously sends a backplane TTL Trigger to the other two modules and goes to its next List point.  
The second module receives the backplane TTL Trigger and immediately goes to its next List point.  
The third module receives the backplane TTL Trigger, immediately goes to its next List point.  
The computer gets a measurement from the measurement instrument.  
90 Application Programs  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MPS Set Up  
Module in slot 0:  
The module is connected to + 5 V on the DUT.  
The initial voltage setting is 0 V.  
Set the voltage mode to List.  
Download the voltage List.  
Set the List to be trigger paced.  
The module listens for the computer to send a trigger command. Upon receipt of the trigger command, the module outputs  
its next List point.  
Also upon receipt of the trigger command, the module generates a backplane TTL Trigger.  
Module in slot 1:  
The module is connected to + 12 V on the DUT.  
The initial voltage setting is 0 V.  
Set the voltage mode to List.  
Download the voltage List.  
Set the List to be trigger paced.  
The module listens for a backplane TTL Trigger.  
Upon receipt of a trigger, the module goes to its next List point.  
Module in slot 2.  
The module is connected to - 12 V on the DUT.  
The initial voltage setting is 0 V.  
Set the voltage mode to List.  
Download the voltage List.  
Set the List to be trigger paced,  
The module listens for a backplane TTL Trigger.  
Upon receipt of a trigger, the module goes to its next List point.  
Variations On This Implementation  
1. A current List could also have been executed by the module so that for each voltage point, a corresponding current limit  
could be programmed.  
2. Overcurrent protection could be enabled to protect a faulty DUT.  
3. The module could generate an SRQ when it finishes changing voltage for each point in the List based on the STC (Step  
Completed) status bit, which indicates when the module has completed executing the next point in the List. The SRQ  
could tell the computer to get a measurement from the measurement instrument.  
4. The module could be told to output its next List point in response to an external or backplane TTL Trigger. (see next  
application.)  
Application Programs 91  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Figure B6-1. Block Diagram of Application #6  
Figure B6-2. Timing Diagram of Application #6  
92 Application Programs  
Download from Www.Somanuals.com. All Manuals Search And Download.  
10  
! APPLICATION #6: OUTPUT SEQUENCING PACED BY THE COMPUTER  
20  
! PROGRAM: APP-6  
30  
!
40  
50  
DIM Plus_5v$[50],Plus_12v$[50],Minus_12v$[50]  
!
60  
70  
80  
Plus_5v$=”4.75, 5, 5.25, 5, 5, 5, 5”  
Ptus_12v$="12, 12, 12, 11.4, 12.6, 12, 12”  
Minus_12v$="12, 12, 12, 12, 12, 11.4, 12.6”  
!
!
!
THESE ARE THE BIAS  
SUPPLY LIMIT CONDITIONS  
TO BE TESTED  
90  
!
100  
110  
120  
130  
140  
150  
160  
170  
180  
190  
200  
210  
220  
230  
240  
250  
260  
270  
280  
290  
300  
310  
320  
330  
340  
350  
360  
370  
380  
390  
400  
410  
420  
430  
440  
450  
460  
470  
480  
490  
500  
510  
520  
530  
540  
550  
560  
570  
580  
590  
Num_test_steps=7  
Dwell=.0I0  
!
ASSIGN @Slot0 TO 70500  
ASSIGN @Slot1 TO 70501  
ASSIGN @Slot2 TO 70502  
!
! SET UP MODULE IN SLOT 0 AS +5 V BIAS SUPPLY  
!
OUTPUT @Slot0;"*RST;*CLS;STATUS:PRESET"  
OUTPUT @Slot0;”VOLT 0”  
!
!
NUMBER OF BIAS SUPPLY LIMIT C0MBINATIONS  
SECONDS OF DWELL TIME  
! SELECT CODE 7, MAINFRAME ADDRESS 05, SLOT 00  
! SELECT CODE 7, MAINFRAME ADDRESS 05, SLOT 01  
! SELECT CODE 7, MAINFRAME ADDRESS 05, SLOT 02  
---------------------  
!
RESET AND CLEAR MODULE  
START AT 0 V  
ENABLE OUTPUT  
!
!
!
!
!
!
!
OUTPUT @Slot0;”OUTPUT ON"  
OUTPUT @Slot0;”VOLTAGE:MODE LIST"  
OUTPUT @Slot0;”LIST:VOLTAGE ";Plus_5v$  
OUTPUT @Slot0;”LIST:DWELL”;Dwell  
OUTPUT @Slot0;”LIST:STEP ONCE"  
OIJTPUT @Slot0;”TRIGGER:SOURCE BUS"  
OUTPUT @Slot0;”OUTPUT:TTLTRG:SOURCE BUS"!  
OUTPUT @Slot0;"OUTPUT:TTLTRG:STATE ON"  
OUTPUT @Slot0;”INITIATE"  
!
! SET UP MODULE IN SLOT 1 AS +12 V BIAS SUPPLY  
!
SET TO GET VOLTAGE FROM LIST  
DOWNLOAD VOLTAGE LIST POINTS  
DOWNLOAD I DWELL TIME (ASSUMES SAME FOR ALL POINTS)  
EXECUTE 1 LIST POINT PER TRIGGER  
TRIGGER SOURCE IS GPIB 'BUS'  
GENERATE BACKPLANE TTL TRIGGER WHEN GPIB 'BUS' TRIGGER IS RECEIVED  
ENABLE TTL TRIGGER DRIVE  
ENABLE RESPONSE TO TRIGGER  
!
!
---------------------  
OUTPUT @Slot1;"*RST;*CLS;STATUS:PRESET"  
OUTPUT @Slot1;"VOLT 0”  
!
!
!
!
!
!
!
!
!
RESET AND CLEAR MODULE  
START AT 0 V  
ENABLE OUTPUT  
SET TO GET VOLTAGE FROM LIST  
DOWNLOAD VOLTAGE LIST POINTS  
DOWNLOAD 1 DWELL TIME (ASSUMES SAME FOR ALL POINTS)  
EXECUTE 1 LIST POINT PER TRIGGER  
TRIGGER SOURCE IS BACKPLANE TTL TRIGGER  
ENABLE RESPONSE TO TRIGGER  
OUTPUT @Slot1;"OUTPUT ON"  
OUTPUT @Slot1;"VOLT:MODE LIST"  
OUTPUT @Slot1;"LIST:VOLTAGE ";Plus_12v$  
OUTPUT @Slot1;"LIST:DWELL”;Dwell  
OUTPUT @Slot1;"LIST:STEP ONCE"  
OUTPUT @Slot1;"TRIGGER:SOURCE TTLTRG"  
OUTPUT @Slot1;"INITIATE"  
!
! SET UP MODULE IN SLOT 2 AS -12 V BIAS SUPPLY  
!
OUTPUT @Slot2;"*RST;*CLS;STATUS:PRESET"  
OUTPUT @Slot2;"VOLT 0  
OUTPUT @Slot2;"OUTPUT ON"  
OUTPUT @Slot2;"VOLT:MODE LIST"  
OUTPUT @Slot2;"LIST:VOLTAGE ";Minus_12v$  
OUTPUT @Slot2;"LIST:DWELL”;Dwell  
OUTPUT @Slot2;"LIST:STEP ONCE"  
OUTPUT @Slot2;"TRIGGER:SOURCE TTLTRG"  
OUTPUT @Slot2;"INITIATE"  
----------------------  
!
!
!
!
!
!
!
!
!
RESET AND CLEAR MODULE  
START AT 0 V  
ENABLE OUTPUT  
SET TO GET VOLTAGE FROM LIST  
DOWNLOAD VOLTAGE LIST POINTS  
DOWNLOAD 1 DWELL TIME (ASSUMES SAME FOR ALL POINTS)  
EXECUTE 1 LIST POINT PER TRIGGER  
TRIGGER SOURCE IS BACKPLANE TTL TRIGGER  
ENABLE RESPONSE TO TTL TRIGGER  
!
! BEFORE TRIGGERING THE MODULES, DETERMINE IF THE MODULES ARE READY BY CHECKING FOR  
! 'WAITING FOR TRIGGER' (BIT 5 OF THE OPERATION STATUS REGISTER). IF THE LAST MODULE PROGRAMMED  
! IS READY THEN SO ARE THE OTHERS, SO JUST CHECK SLOT 2.  
!
! YOU COULD ELIMINATE THIS STEP BY SIMPLY INSERTING A PAUSE IN THE PROGRAM. HOWEVER, BY  
Application Programs 93  
Download from Www.Somanuals.com. All Manuals Search And Download.  
600  
610  
620  
630  
640  
650  
660  
670  
680  
690  
700  
710  
720  
730  
740  
750  
760  
770  
780  
790  
800  
810  
820  
830  
840  
850  
860  
! CHECKING THE INSTRUMENT STATUS, YOU CAN AVOID TIMING PROBLEMS. ALSO, ANY OTHER OPERATIONS  
! THAT TAKE TIME WILL GIVE THE MODULES A CHANCE TO COMPLETE PROCESSING.  
!
REPEAT  
OUTPUT @Slot2;"STATUS:OPERATION:CONDITION?"  
ENTER @Slot2;Condition_data  
UNTIL BIT(Condition_data,5)  
!
!
TEST FOR BIT 5 = TRUE  
! GENERATE A TRIGGER AND MAKE A MEASUREMENT FOR EACH TEST CONDITION  
!
FOR Loop_count=1 TO Num_test_steps  
OUTPUT @Slot0;"*TRG"  
!
SEND Agilent -1B BUS TRIGGER  
GOSUB Get_measurement  
NEXT Loop_count  
!
STOP  
!
Get_measurement:  
!
! THIS IS JUST TO SHOW YOU WHERE YOU WOULD ADD CODE TO GET DATA FROM THE MEASUREMENT INSTRUMENT.  
! THE MEASUREMENT MUST TAKE LONGER THAN THE PROGRAMMED DWELL TIME OR YOU WILL MISS TRIGGERS.  
!
WAIT .1  
!
RETURN  
!
END  
Figure B6-3. Agilent BASIC Program Listing for Application #6  
94 Application Programs  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Application 7. Output Sequencing Without Computer Intervention  
Overview Of Application  
When characterizing devices, the DUT’s performance is measured over a range of power supply voltages. This test can be  
performed without computer intervention by using hardware signals from the measurement instrument to cause the power  
module to sequence to the next voltage in a preprogrammed List. By buffering these readings in the measurement  
instrument, the entire test can be executed without computer involvement. For characterizations that require long  
measurement times, the computer is free to do other tasks. For characterizations that must execute at hardware speeds, the  
computer is not involved and will not slow down the test loop.  
In this example, the power module must apply 8 to 14 volts (in 13 0.5-volt increments) to an automotive engine sensor. The  
module varies the bias voltage to the engine sensor and the sensor’s output is measured to characterize its performance over  
the range of possible "battery voltages". The sensor output is measured by a DMM that has an internal buffer and stores  
each reading.  
By combining Lists and trigger capabilities, the MPS can be used to address this application. The module can be  
programmed to use its triggering capabilities to the fullest extent. Each time the module executes the next step in its List  
and changes voltage, the module will generate an external trigger. The external trigger will cause the DMM, equipped with  
an external trigger input, to take and store a reading. The DMM, also equipped with a "Measurement Complete" output,  
sends its output trigger signal to the module to tell the module to go to its next List point. Effectively, the module and the  
DMM "handshake", so that the two function at hardware speeds without computer intervention.  
When the test is complete, either device can signal the computer to get the data from the DMM. For the purpose of this  
example, the module will generate an SRQ when the last List point has been executed. This is indicated by the OPC  
(Operation Complete) bit in the status register.  
Another detail that needs attention is timing. The DUT may require some settling time before the DMM is told to take a  
reading. The module’s dwell time can be used to do this. The STC (Step Complete) status signal indicates when the point  
has been executed and its dwell time has expired. The dwell time is programmed to be the engine sensor’s settling time.  
The external trigger is generated when STC is asserted. Thus, the DMM will not be triggered until the dwell time has  
expired and the sensor’s output has settled.  
This type of self-paced test execution is useful in two situations. When the test must execute very fast, there is no time for  
the computer to be involved in each iteration of the test loop. Therefore, the test must execute without computer  
intervention. The second situation is when the test is very long. For example, if the measurement instrument took 1 minute  
to make each measurement, the test would take 13 minutes to execute. The computer is not used efficiently if it is idle while  
waiting for each measurement loop, so it would be best to have the computer executing another task. Without self-pacing,  
you would need to develop interrupt driven software that stops every 1 minute to take a reading. By letting the module and  
the DMM run on their own, code development is much simpler and computer resources are used more efficiently.  
MPS Features Used  
20-point voltage List.  
Dwell time.  
Trigger-paced Lists.  
Generate an SRQ on a change in internal status.  
Generate a trigger on a change in internal status.  
Trigger in/out from MPS mainframe backplane TTL Trigger.  
Trigger on a GPIB trigger command.  
Application Programs 95  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Advantages/Benefits Of The MPS Solution  
The entire test executes without computer involvement, the command processing time is eliminated from the test loop.  
The entire test executes without computer involvement, so the computer can perform other tasks while the test executes.  
Software development is simplified; you do not need to write a test loop because the module and the DMM are running on  
their own.  
By using dwell times, the trigger out signal can be sent at the correct time, which permits the DUT to settle before a reading  
is taken.  
Implementation Details  
How The MPS Implements The Sequence  
The module listens for the computer to send a trigger command.  
Upon receipt of the trigger command, the module outputs its first List point.  
After the dwell time expires, the STC is asserted and the module generates an external trigger.  
The DMM receives the external trigger, takes and stores a reading.  
The DMM generates a "Measurement Complete" output signal when it’s done.  
The module receives the DMM output signal as an external trigger in.  
Also upon receipt of the trigger in, the module outputs its next List point.  
The process repeats for each List point.  
After the last List point has executed, the module generates SRQ, telling the computer the test has completed.  
MPS Set Up  
Set the voltage mode to List.  
Download the voltage List.  
Download the dwell time List.  
Set the List to be trigger paced.  
Set the trigger source to external trigger.  
Note  
The computer trigger command "TRIGGER:IMMEDIATE" is always active, even if the external trigger  
is the selected source.  
Set the module to generate a backplane TTL Trigger on STC. This backplane TTL Trigger drives external trigger out.  
Set the module to generate SRQ on OPC.  
Variations On This Implementation  
1. A current List could also have been executed by the module so that for each voltage point, a corresponding current limit  
could be programmed.  
2. Overcurrent protection could be enabled to protect a faulty engine sensor.  
3. If the DMM does not have an internal buffer, the computer could take a reading on each iteration of the test loop (see  
previous application).  
96 Application Programs  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Figure B7-1. Block Diagram of Application #7  
Figure B7-2. Timing Diagram of Application #7  
Application Programs 97  
Download from Www.Somanuals.com. All Manuals Search And Download.  
10  
! APPLICATION #7: OIJTPUT SEQUENCING WITHOUT COMPUTER INTERVENTION  
20  
30  
! PROGRAM: APP_7  
!
40  
ASSIGN @Slot0 TO 70500  
!
SELECT CODE 7, MAINFRAME ADDRESS 05, SLOT 00  
50  
!
60  
DIM Vlist$[80]  
70  
80  
Vlist$=”8, 8.5, 9, 9.5, 10, 10.5, 11, 11.5, 12, 12.5, 13, 13.5, 14” ! VOLTAGE LIST POINTS  
!
90  
OUTPUT @Slot0;"*RST;*CLS;STATUS:PRESET"  
OUTPUT @Slot0;”VOLT 0”  
OUTPUT @Slot0;”CURR 1"  
OUTPUT @Slot0;"OUTPUT ON"  
OUTPUT @Slot0;”VOLT:MODE LIST"  
OUTPUT @Slot0;”LIST:VOLT ";Vlist$  
OUTPUT @Slot0;”LIST:DWELL .050”  
!
!
!
!
!
!
!
!
!
RESET AND CLEAR MODULE  
START AT 0 V  
SET CURRENT LIMIT  
ENABLE OUTPUT  
SET TO GET VOLTAGE FROM LIST  
DOWNLOAD VOLTAGE LIST POINTS  
DOWNLOAD 1 DWELL POINT (ASSUMES SAME FOR ALL POINTS)  
USE A 50 ms SETTLING TIME AS THE DWELL TIME  
EXECUTE 1 POINT PER TRIGGER  
100  
110  
120  
130  
140  
150  
160  
170  
180  
190  
200  
210  
220  
230  
240  
250  
260  
270  
280  
290  
300  
310  
320  
330  
340  
350  
360  
370  
380  
390  
400  
410  
420  
430  
440  
450  
460  
470  
480  
490  
500  
510  
520  
530  
540  
550  
560  
570  
580  
590  
OUTPUT @Slot0;”LIST:STEP ONCE"  
!
OUTPUT @Slot0;"*ESE 1”  
!
!
!
!
ENABLES DETECTION OF OPC IN THE STANDARD EVENT REGISTER.  
OPC = BIT 0 = VALUE 1 OF THE STANDARD EVENT REGISTER.  
ENABLES THE SERVICE REQUEST REGISTER TO GENERATE AN SRQ WHEN  
ANY EVENT IN THE STANDARD EVENT REGISTER IS ASSERTED.  
THE STANDARD EVENT REGISTER = BIT 5 = VALUE 32.  
OUTPUT @Slot0;"*SRE 32”  
!
OUTPUT @Slot0;"OUTPUT:TTLTRG:STATE ON"  
OUTPUT @Slot0;”OUTPUT:TTLTRG:SOURCE LINK"  
OUTPUT @Slot0;”OUTPUT:TTLTRG:LINK ‘STC’”  
OUTPUT @Slot0;”TRIGGER:SOURCE EXTERNAL" !  
OUTPUT @Slot0;’INITIATE"  
OUTPUT @Slot0;"*OPC”  
!
ENABLE BACKPLANE TTL TRIGGER DRIVE  
!
WHEN THE MODULE INDICATES SIC (STEP COMPLETED),  
!
GENERATE A BACKPLANE TTL TRIGGER  
USE EXTERNAL TRIGGER IN BNC AS TRIGGER SOURCE  
ENABLE RESPONSE TO TRIGGER  
TELLS MODULE TO ASSERT OPC (OPERATION COMPLETE)  
WHEN IT COMPLETES THE LIST. OPC GENERATES SRO.  
!
!
!
!
ON INTR.7 GOSUB Srq_handler  
ENABLE INTR 7;2  
!
! BEFORE TRIGGERING THE MODULE, DETERMINE IF IT IS READY BY CHECKING FOR  
! 'WAITING FOR TRIGGER' (BIT 5 OF THE OPERATION STATUS REGISTER).  
!
!
ENABLE SRQ INTERRUPT AND  
IDENTIFY HANDLER SUBROUTINE  
!
! YOU COULD ELIMINATE THIS STEP BY SIMPLY INSERTING A PAUSE IN THE PROGRAM. HOWEVER, BY  
! CHECKING THE INSTRUMENT STATUS, YOU CAN AVOID TIMING PROBLEMS. ALSO, ANY OTHER OPERATIONS  
! THAT TAKE TIME WILL GIVE THE MODULE A CHANCE TO COMPLETE PROCESSING.  
!
REPEAT  
OUTPUT @Slot0;”STATUS:OPERATION:CONDITION?"  
ENTER @Slot0;Condition_data  
UNTIL BIT(Condition_data,5)  
!
! BEGIN THE SELF-PACED TEST LOOP BY TRIGGERING THE MODULE TO START THE LIST  
!
OUTPUT @Slot0;”TRIGGER:IMMEDIATE"  
!
TEST FOR BIT 5 = TRUE  
!
!
THIS IS AN IMMEDIATE TRIGGER, WHICH IS ALWAYS ACTIVE.  
IT DOES NOT NEED TO BE SELECTED AS TRIGGER SOURCE.  
!
GOTO 530  
!
!
IDLE IN LOOP WAITING FOR SRQ OR GO DO OTHER TASKS  
Srq_handler:  
!
!
! ADD LINES HERE TO READ THE DATA BUFFER FROM THE DMM  
!
END  
Figure B7-3. Agilent BASIC Program Listing of Application #7  
98 Application Programs  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Supplemental Information  
This appendix contains program listings translated into the following DOS-compatible languages and GPIB interfaces:  
GWBASIC and the Agilent 61062/82990/82335A GPIB Command Library for MS-DOS  
GWBASIC and the National Instruments GPIB-PC Interface Card  
Microsoft C and the Agilent 61062/82990/82335A GPIB Command Library for MS-DOS  
Microsoft C and the National Instruments GPIB-PC Interface Card  
Each program is translated from the Agilent BASIC listing found in application #3. This example program was chosen as  
representative of all application programs because it shows how to:  
Configure the interface card.  
Address the power module.  
Write strings to the power module.  
Write real arrays to the power module.  
Receive real numbers from the power module.  
The six other application programs all use a subset of the above functions.  
1
MERGE "SETUP.BAS" AS DESCRIBED IN YOUR GPIB COMMAND LIBRARY MANUAL  
2
1000  
1010  
1020  
1030  
1040  
1050  
1060  
1070  
1080  
1090  
1100  
1110  
1120  
1130  
1140  
1150  
1160  
1170  
1180  
1190  
1200  
1210  
1220  
1230  
1240  
1250  
1260  
1270  
1280  
1290  
1300  
1310  
1320  
1330  
APPLICATION #3: CONTROLLING VOLTAGE RAMP UP AT TURN ON  
FOR GWBASIC AND THE Agilent 61062/82990/82335A GPIB COMMAND LIBRARY  
PROGRAM: Agilent 3.BAS  
OPTION BASE 1  
INTERFACE = 7  
SLOTO 705001  
CR.LFS CHR$(13) + CHR$(10)  
DIM VSTEP(20)  
NUM.POINTS = 20  
VSTART = 2  
SELECT CODE OF THE GPIB CARD  
SELECT CODE 7, MAINFRAME ADDRESS 05, SLOT 00  
CARRIAGE RETURN + LINE FEED = END OF LINE TERMINATION  
ARRAY TO HOLD THE VOLTAGE RAMP STEPS  
NUMBER OF POINTS IN THE VOLTAGE RAMP ARRAY  
START VOLTAGE FOR RAMP  
VSTOP = 10  
STOP VOLTAGE FOR RAMP  
RAMPTIME = .5  
DWELL = RAMPTIME / 19  
TIME IN SECONDS TO CHANGE FROM VSTART TO VSTOP  
DWELL TIME FOR EACH POINT  
SINCE THE OUTPUT STAYS AT THE LAST VOLTAGE POINT AFTER ITS DWELL TIME EXPIRES, THE DWELL TIME OF THE  
LAST POINT IS NOT PART OF THE TRANSITION TIME. THEREFORE, DIVIDE THE TOTAL TIME BY 19 POINTS, NOT 20.  
YOU WANT THE SAME DWELL TIME FOR EVERY POINT IN THE LIST, SO YOU NEED TO DOWNLOAD ONLY 1 DWELL TIME.  
FOR I=1 TO 20  
VSTEP(I) = VSTART + ((( VSTOP - VSTART ) / 20 ) * I )  
NEXT I  
CALCULATES VOLTAGE LIST POINTS  
NOTE REGARDING GPIB READ/WRITE TERMINATIONS:  
THE DEFAULT MODE OF THE INTERFACE CARD IS THAT EOI IS ENABLED AND THE READ/WRITES TERMINATE  
ON CARRIAGE RETURN/LINE FEED. THE MODULE TERMINATES ON EITHER EOI OR LINE FEED, SOTHE  
DEFAULT SETTINGS OF THE CARD ARE SUFFICIENT.  
CMD$ = "*RST;*CLS;STATUS:PRESET"  
L = LEN( CMD$ )  
RESET AND CLEAR MODULE  
CALL IOOUTPUTS( SLOTO, CMD$, L )  
IF PCIB.ERR<>0 THEN ERROR PCIB.BASERR  
Application Programs 99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1340  
1350  
1360  
1370  
1380  
1390  
1400  
1410  
1420  
1430  
1440  
1450  
1460  
1470  
1480  
1490  
1500  
1510  
1520  
1530  
1540  
1550  
1560  
1570  
1580  
1590  
1600  
1610  
1620  
1630  
1640  
1650  
1660  
1670  
1680  
1690  
1700  
1710  
1720  
1730  
1740  
1750  
1760  
1770  
1780  
1790  
1800  
1810  
1820  
1830  
1840  
1850  
1860  
1870  
1880  
1890  
1900  
1910  
1920  
1930  
CMD$ = "VOLT+ STR$( VSTART )  
L = LEN( CMD$ )  
CALL IOOUTPUTS( SLOTO, CMD$, L )  
IF PCIB.ERR<>0 THEN ERROR PCIB.BASERR  
CMD$ = "CURR .1”  
L = LEN( CMD$ )  
CALL IOOUTPUTS( SLOTO, CMD$, L )  
IF PCIB.ERR<>0 THEN ERROR PCIB.BASERR  
START RAMP AT VSTART. USE NUMBER TO STRING  
CONVERSION TO SEND REAL NUMBERS OVER THE BUS  
AS PART OF THE COMMAND STRING.  
CMD$ = "OUTPUT ON"  
L = LEN( CMD$ )  
ENABLE OUTPUT  
CALL IOOUTPUTS( SLOTO, CMD$, L )  
IF PCIB.ERR<>0 THEN ERROR PCIB.BASERR  
CMD$ = "VOLT:MODE LIST"  
L = LEN( CMD$ )  
CALL IOOUTPUTS( SLOTO, CMD$, L )  
IF PCIB.ERR<>0 THEN ERROR PCIB.BASERR  
SENDING THE VOLTAGE DATA POINTS REQUIRES TWO STEPS USING THE GPIB COMMAND LIBRARY. THE INSTRUCTION CONTAINS BOTH  
STRING DATA AND A REAL ARRAY. FIRST, SEND THE STRING DATA COMMAND HEADER "LIST:VOLT" TO THE MODULE USING lOOUTPUTS.  
THEN, SEND THE REAL ARRAY USING IOOUTPUTA. HOWEVER, YOU MUST INHIBIT THE EOI AND END-OF-LINE TERMINATOR AFTER THE  
‘ IOOUTPUTS COMMAND OR THE MODULE WILL STOP TAKING DATA. THEN RE-ENABLE THEM TO TERMINATE THE IOOUTPUTA.  
SET TO GET VOLTAGE FROM LIST  
EOI.STATE = 0  
CALL IOEOI( INTERFACE, EOI.STATE )  
TURN OFF EOI  
IF PCIB.ERR<>0 THEN ERROR PCIB.BASERR  
END.OF.LINE = 0  
CALL IOEOL( INTERFACE, CR.LF$, END.OF.LINE )  
TURN OFF END-OF-LINE TERMINATION  
IF PCIB.ERR<>0 THEN ERROR PCIB.BASERR  
CMD$ = "LIST:VOLT  
L = LEN( CMD$ )  
CALL IOOUTPUTS( SLOTO, CMD$, L )  
SEND THE VOLTAGE HEADER (STRING) ...  
TURN ON EOI  
IF PCIB.ERR<>0 THEN ERROR PCIB.BASERR  
EOI.STATE = 1  
CALL IOEOI( INTERFACE, EOI.STATE )  
IF PCIB.ERR<>0 THEN ERROR PCIB.BASERR  
END.OF.LINE = LEN (CR.LF$ )  
CALL IOEOL( INTERFACE, CR.LF$ , END.OF.LINE )  
IF PCIB.ERR<>0 THEN ERROR PCIB.BASERR  
CALL IOOUTPUTA( SLOTO, VSTEP(1), NUM.POINTS ) ‘ DOWNLOAD THE VOLTAGE POINTS (ARRAY)  
IF PCIB.ERR<>0 THEN ERROR PCIS.BASERR  
CMD$ = "LIST:DWELL “ + STR$( DWELL )  
L = LEN( CMD$ )  
CALL IOOUTPUTS( SLOTO, CMD$, L )  
IF PCIB.ERR<>0 THEN ERROR PCIB.BASERR  
CMD$ = "LIST:STEP AUTO"  
L = LEN( CMD$ )  
CALL IOOUTPUTS( SLOTO, CMD$, L )  
IF PCIB.ERR<>0 THEN ERROR PCIB.BASERR  
TURN ON END-OF-LINE TERMINATION  
DOWNLOAD 1 DWELL TIME. USE NUMBER TO STRING  
CONVERSION TO SEND REAL NUMBERS OVER THE BUS  
AS PART OF THE COMMAND STRING.  
DWELL-PACED LIST  
CMD$ = "INITIATE"  
ENABLE TRIGGER TO START LIST  
100 Application Programs  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1940  
1950  
1960  
1970  
1980  
1990  
2000  
2010  
2020  
2030  
2040  
2050  
2060  
2070  
2080  
2090  
2100  
2110  
2120  
2130  
2140  
2150  
2160  
2170  
2180  
2190  
2200  
2210  
2220  
L = LEN( CMD$ )  
CALL IOOUTPUTS( SLOTO, CMDS, L )  
IF PCIB.ERR<>0 THEN ERROR PCIB.BASERR  
BEFORE TRIGGERING THE MODULE, DETERMINE IF IT IS READY BY CHECKING FOR  
' WAITING FOR TRIGGER' (BIT 5 OF THE OPERATION STATUS REGISTER).  
YOU COULD ELIMINATE THIS STEP BY SIMPLY INSERTING A PAUSE IN THE PROGRAM. HOWEVER, BY  
CHECKING THE INSTRUMENT STATUS, YOU CAN AVOID TIMING PROBLEMS. ALSO, ANY OTHER OPERATIONS  
THAT TAKE TIME WILL GIVE THE MODULE A CHANCE TO COMPLETE PROCESSING.  
CONDITION.DATA = 0  
WHILE ( ( CONDITION.DATA AND 32 ) ) <> 32 )  
CMD$ = "STATUS:OPERATION:CONDITION?"  
L = LEN( CMD$ )  
CALL IOOUTPUTS( SLOTO, CMD$, L )  
IF PCIB.ERR<>0 THEN ERROR PCIB.BASERR  
CALL IOENTER( SLOTO, CONDITION.DATA )  
IF PCIB.ERR<>0 THEN ERROR PCIB.BASERR  
WEND  
CONTINUE TO LOOP UNTIL BIT 5 (VALUE 32) = TRUE  
SEND TRIGGER COMMAND TO START LIST AND GENERATE THE VOLTAGE RAMP  
CMD$ = "TRIGGER:IMMEDIATE"  
L = LEN( CMD$ )  
CALL IOOUTPUTS( SLOTO, CMD$, L )  
IF PCIB.ERR<>0 THEN ERROR PCIB.BASERR  
THIS IS AN IMMEDIATE TRIGGER, WHICH IS ALWAYS  
ACTIVE. THEREFORE, IT DOES NOT NEED TO BE  
SELECTED AS A TRIGGER SOURCE.  
2230 END  
Application Programs 101  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1
MERGE "DECL.BAS" AS INSTRUCTED IN YOUR NATIONAL INSTRUMENTS GPIB-PC MANUAL  
2
1000  
1010  
1020  
1030  
1040  
1050  
1060  
1070  
1080  
1090  
1100  
1110  
1120  
1130  
1140  
1150  
1160  
1170  
1180  
1190  
1200  
1210  
1220  
1230  
1240  
1250  
1260  
1270  
1280  
1290  
1300  
1310  
1320  
1330  
1340  
1350  
1360  
1370  
1380  
1390  
1400  
1410  
1420  
1430  
1440  
1450  
1460  
1470  
1480  
1490  
1500  
1510  
1520  
1530  
1540  
1550  
1560  
1570  
APPLICATION #3: CONTROLLING VOLTAGE RAMP UP AT TURN ON  
FOR GWBASIC AND THE NATIONAL INSTRUMENTS GPIB-PC INTERFACE CARD  
PROGRAM: N3.BAS  
CONFIGURE THE GPIB.COM HANDLER FOR THE FOLLOWING:  
EOI ENABLED FOR BOTH READ AND WRITE  
DISABLE AUTO SERIAL POLL  
‘ INSTRUMENT.NAME$ = "SLOTO"  
CALL IBFIND( INSTRUMENT.NAME$, SLOTO% )  
IF SLOTO% < 0 THEN PRINT “COULDN'T FIND MODULE" : STOP  
OPTION BASE 1  
VSTEP$ = “”  
VSTART = 2  
VSTOP = 10  
RAMPTIME = .5  
DWELL = RAMPTIME / 19  
STRING TO HOLD THE VOLTAGE RAMP STEPS  
START VOLTAGE FOR RAMP  
STOP VOLTAGE FOR RAMP  
TIME IN SECONDS TO CHANGE FROM VSTART TO VSTOP  
DWELL TIME FOR EACH POINT  
SINCE THE OUTPUT STAYS AT THE LAST VOLTAGE POINT AFTER ITS DWELL TIME EXPIRES, THE DWELL TIME OF THE  
' LAST POINT IS NOT PART OF THE TRANSITION TIME. THEREFORE, DIVIDE THE TOTAL TIME BY 19 POINTS, NOT 20.  
' YOU WANT THE SAME DWELL TIME FOR EVERY POINT IN THE LIST, SO YOU NEED TO DOWNLOAD ONLY 1 DWELL TIME.  
'
' SINCE THE NATIONAL INSTRUMENTS GPIB-PC WORKS WITH STRINGS, THE RAMP DATA MUST BE CONSOLIDATED INTO A  
' SINGLE STRING WHICH CONTAINS ALL THE POINTS, SEPARATED BY COMMAS.  
FOR I=1 TO 20  
MAKES THE STRING EQUIVALENTS OF THE  
VSTEP$ = VSTEP$ + STR$( VSTART + ((( VSTOP - VSTART ) / 20 *I )) ‘ VOLTAGE POINTS AND CONCATENATES THEM ONLY  
IF I <> 20 THEN VSTEP$=VSTEP$+","  
NEXT I  
FOR THE FIRST 19 POINTS, EACH FOLLOWED  
BY A COMMA. THE LAST POINT IS NOT  
FOLLOWED BY A COMMA.  
CMD$ = "*RST;*CLS;STATUS:PRESET"  
CALL IBWRT( SLOTO%, CMD$ )  
IF IBSTA% < 0 THEN GOTO 1960  
RESET AND CLEAR MODULE  
CMD$ = "VOLT “ + STR$( VSTART )  
CALL IBWRT( SLOTO%, CMD$ )  
IF IBSTA% < 0 THEN GOTO 1960  
START RAMP AT VSTART. USE NUMBER TO STRING  
CONVERSION TO SEND REAL NUMBERS OVER THE BUS  
AS PART OF THE COMMAND STRING.  
CMD$  
= "CURR .1"  
CALL IBWRT( SLOTO%, CMD$  
IF IBSTA% < 0 THEN GOTO 1960  
CMD$ = "OUTPUT ON"  
CALL IBWRT( SLOTO%, CMD$ )  
IF IBSTA% < 0 THEN GOTO 1960  
ENABLE OUTPUT  
CMD$ = "VOLT:MODE LIST"  
CALL IBWRT( SLOTO%, CMD$ )  
IF IBSTA% < 0 THEN GOTO 1960  
CMD$ = "LIST:VOLT “ + VSTEP$  
CALL IBWRT( SLOTO%, CMD$ )  
IF IBSTA% < 0 THEN GOTO 1960  
SET TO GET VOLTAGE FROM LIST  
DOWNLOAD VOLTAGE LIST POINTS  
DOWNLOAD 1 DWELL TIME. USE NUMBER TO STRING  
CMD$ = "LIST:DWELL” + STR$(DWELL )  
102 Application Programs  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1580  
1590  
1600  
1610  
1620  
1630  
1640  
1650  
1660  
1670  
1680  
1690  
1700  
1710  
1720  
1730  
1740  
1750  
1760  
1770  
1780  
1790  
1800  
1810  
1820  
1830  
1840  
1850  
1860  
1870  
1880  
1890  
1900  
1910  
1920  
1930  
1940  
1950  
1960  
1970  
1980  
1990  
CALL IBWRT( SLOTO%, CMD$ )  
IF IBSTA% < 0 THEN GOTO 1960  
CMD$ = "LIST:STEP AUTO"  
CALL IBWRT( SLOTO%, CMD$ )  
IF IBSTA% < 0 THEN GOTO 1960  
CONVERSION TO SEND REAL NUMBERS OVER THE BUS  
AS PART OF THE COMMAND STRING.  
DWELL-PACED LIST  
CMD$ = "INITIATE"  
ENABLE TRIGGER TO START LIST  
CALL IBWRT( SLOTO%, CMD$ )  
IF IBSTA% < 0 THEN GOTO 1960  
BEFORE TRIGGERING THE MODULE, DETERMINE IF IT IS READY BY CHECKING FOR  
' WAITING FOR TRIGGER' (BIT 5 OF THE OPERATION STATUS REGISTER).  
YOU COULD ELIMINATE THIS STEP BY SIMPLY INSERTING A PAUSE IN THE PROGRAM. HOWEVER, BY  
CHECKING THE INSTRUMENT STATUS, YOU CAN AVOID TIMING PROBLEMS. ALSO, ANYOTHER OPERATIONS  
THAT TAKE TIME WILL GIVE THE MODULE A CHANCE TO COMPLETE PROCESSING.  
CONDITION.DATA$ = SPACE$(20)  
WHILE ( (VAL( CONDITION.DATA$ ) AND 32 ) <> 32) ‘ CONTINUE TO LOOP UNTIL BIT 5 (VALUE 32) = TRUE  
CMD$ = "STATUS:OPERATION:CONDITION?"  
CALL IBWRT( SLOTO%, CMD$ )  
IF IBSTA% < 0 THEN GOTO 1960  
CALL IBRD( SLOTO%, CONDITION.DATA$ )  
IF IBSTA% < 0 THEN GOTO 1960  
WEND  
RESERVE SPACE FOR READING IN STRING  
SEND TRIGGER COMMAND TO START LIST AND GENERATE THE VOLTAGE RAMP  
CMD$ = "TRIGGER:IMMEDIATE"  
CALL IBWRT( SLOTO%, CMDS )  
IF IBSTA% < 0 THEN GOTO 1960  
' THIS IS AN IMMEDIATE TRIGGER, WHICH IS ALWAYS  
' ACTIVE THEREFORE, IT DOES NOT NEED TO BE  
' SELECTED AS A TRIGGER SOURCE.  
STOP  
GENERAL ERROR HANDLER  
PRINT “GPIB function call error: “  
PRINT "IBSTA% =; IBSTA%, "IBERR% =”; IBERR% “lBCNT% = “;IBCNT%  
END  
Application Programs 103  
Download from Www.Somanuals.com. All Manuals Search And Download.  
/* APPLICATION #3: CONTROLLING VOLTAGE RAMP UP AT TURN ON  
FOR MICROSOFT C AND THE Agilent 61062/82990/82335A GPIB COMMAND LIBRARY FOR MS-DOS PROGRAM: Agilent3.C  
#include <stdio.h>  
#include <stdlib.h>  
#include <string.h>  
#include "chpib.h”  
#include "cfunc.h”  
#define INTERFACE 7L  
/* Select code 7 for the Agilent -1B interface card.  
/* Select code 7, mainframe address 05, slot 00. */  
/* Waiting for Trigger (WTG) = bit 5 = value 32. */  
*/  
#define SLOTO  
#define WTG  
70500L  
32  
#define NUM_PTS 20  
/* 20 points in the voltage List.  
*/  
int error;  
main( )  
{
char *cmd;  
char cmd-buff[255];  
static char cr_lf[3] = { 13, 10, 0 };  
int i;  
/* Used to hold command strings sent to the module. */  
/* Used to hold command strings during string manipulations. */  
/* Carriage return+line feed = end of line. */  
/* Loop counter. */  
float condition_data;  
float vstart = 2.0;  
/* Used to hold data from read back of status conditions. */  
/* Start voltage for the ramp. */  
float vstop = 10.0;  
float vstep[NUM_PTS];  
float ramptime = 0.5;  
float dwell;  
/* Stop voltage for the ramp. */  
/* Used to hold voltage List points for the ramp. */  
/* Transition time (in seconds) for the ramp. */  
/* Dwell time (in seconds) for each ramp step. */  
/* Since the output stays at the last voltage point after its dwell  
time expires, the dwelt time of the last point is not part of the transition time.  
Therefore, divide the total time by 19 points, not 20. You want the same dwell  
time for every point in the List, so only download 1 dwell time. */  
dwell = ramptime / 19.0;  
for (i = 1; i <= NUM_PTS; i++)  
/*  
Calculate the voltage List points */  
vstep[i] = vstart + (((vstop - vstart) / NUM_PTS) * i );  
error = ioreset(INTERFACE);  
/* To get the interface to its defaults. */  
/* Enables timeout of 2 seconds. */  
error_handler(error, "Resetting the interface");  
error = iotimeout(INTERFACE, (double)2.0);  
error_handler(error, "Setting the timeout”);  
/* Note regarding GPIB read/write terminations:  
The default of the interface card is that EOI is enabled and the read/writes terminate on carriage return/line feed. The module terminates  
on either EOI or Line feed, so the default settings of the card are sufficient.  
*/  
cmd = "*RST;*CLS;STATUS:PRESET”;  
error = iooutputs(SLOTO, cmd, strlen(cmd));  
error_handler(error, cmd);  
/* Reset and clear module. */  
sprintf(cmd_buff , "VOLT %f”, vstart);  
/* Start ramp at vstart. Use number to string */  
error = iooutputs(SLOTO, cmd_buff, strlen(cmd_buff)); /* conversion to send real numbers over the */  
error_handler(error, cmd-buff);  
/* bus as part of the command string.  
*/  
cmd = "CURR .1”;  
error = iooutputs(SLOT0, cmd, strlen(cmd));  
error-handler(error, cmd);  
cmd = “OUTPUT ON";  
/* Enable output */  
104 Application Programs  
Download from Www.Somanuals.com. All Manuals Search And Download.  
error = iooutputs(SLOTO, cmd, strlen(cmd));  
error_handler(error, cmd);  
cmd = "VOLT:MODE LIST”;  
/* Set to get voltage from List */  
error = iooutputs(SLOTO, cmd, strlen(cmd));  
error_ handler(error, cmd);  
/* Sending voltage data points requires two steps using the Agilent -1B Command Library. The instruction contains both string data and a  
real  
array. First, send the string data command header "LIST:VOLT “ to the module using iooutputs. Then, send the real array using iooutputa.  
However, you must inhibit the EOI and End-of-Line terminator after the iooutputs or the module will stop taking data. Then, re-enable them  
to terminate the iooutputa.  
*/  
error = ioeoi(INTERFACE, 0);  
/* Turn off EOI */  
error_handler(error, "Disabling EOI”);  
error = ioeol(INTERFACE, “”, 0);  
/* Turn off End-of-Line termination */  
/* First send the voltage header (string) ... */  
error-handler(error, "Disabling EOI”);  
cmd = "LIST:VOLT “;  
error = iooutputs(SLOTO, cmd, strlen(cmd));  
error_handler(error, cmd);  
error = ioeoi(INTERFACE, 1);  
/* Turn on EOI */  
error_handler(error, "Enabling EOI”);  
error = ioeoL(INTERFACE, cr_lf, strlen(cr_lf));  
error_handler(error, "Enabling EOL”);  
/* Turn on End-of-Line termination */  
error = iooutputa(SLOTO, &vstep[1], NUM_PTS);  
error_handler(error, "Voltage List Array");  
/* Download voltage points (array), starting */  
/* with the element 1, not 0. */  
sprintf(cmd_buff, "LIST:DWELL %f”, dwell);  
/* Download 1 dwell time. Use number to */  
error = iooutputs(SLOTO, cmd_buff, strlen(cmd_buff)); /* string conversion to send the real  
*/  
*/  
*/  
error_handler(error, cmd_buff);  
/* number over the bus as part of the  
/* command string.  
cmd = "LIST:STEP AUTO";  
/* Dwell-paced List */  
error = iooutputs(SLOTO, cmd, strlen(cmd));  
error_handler(error, cmd);  
cmd = "INITIATE";  
/* Enable trigger to start List  
error = iooutputs(SLOTO, cmd, strlen(cmd));  
error_handler(error, cmd);  
/* Before triggering the module, determine if it is ready by checking for  
'Waiting for Trigger' (bit 5 of the Operation Status Register).  
You could eliminate this step by simply inserting a pause in the program. However, by checking the instrument status,  
you can avoid timing problems. Also, any other operations that take time will give the module a chance to complete processing. */  
do {  
cmd = "STATUS:OPERATION:CONDITION?”;  
error = iooutputs(SLOTO, cmd, strlen(cmd));  
error_handler(error, cmd);  
error = ioenter(SLOTO, &condition_data);  
error_handler(error, "Readback of status");  
/* You must convert float to integer */  
/* to do an integer bit test. */  
Application Programs 105  
Download from Www.Somanuals.com. All Manuals Search And Download.  
} while (((int)condition_data && WTG) == 0) ;  
/* Loop until bit 5 (value 32) is true. */  
/* Send trigger command to start List and generate the voltage ramp. */  
cmd = "TRIGGER:IMMEDIATE”;  
error = iooutputs(SLOTO, cmd, strlen(cmd));  
error_handler(error, cmd);  
/* This is an immediate trigger, which is always */  
/* active. Therefore, it does not need to be */  
/* selected as a trigger source. */  
}
error_handler(error,bad_string)  
/* This is a generalized error checking routine. */  
int error;  
char *bad_string;  
{
if (error != 0) {  
printf("Agilent -1B error while sending or receiving ‘%s’.\n”, bad_string);  
printf("Error #: %d - - %s\n”, error, errstr(error));  
}
}
106 Application Programs  
Download from Www.Somanuals.com. All Manuals Search And Download.  
/* APPLICATION #3: CONTROLLING VOLTAGE RAMP UP AT TURN ON,  
FOR MICROSOFT C AND THE NATIONAL INSTRUMENTS GPIB-PC INTERFACE CARD  
PROGRAM: N3.C  
Configure the GPIB.COM handler for the following:  
#include <stdio.h>  
EOl enabled for both read and write  
Disable auto serial poll  
*/  
#include <stdlib.h>  
#include <string.h>  
#include "decl.h”  
#define ERR  
#define NUM_PTS 20  
(1<<15)  
/* Error detected as bit 15 of ibsta.  
/* The number of points in the voltage List.  
*/  
*/  
#define MAX_LEN 255  
#define SMALL_STRING 15  
#define WTG 32  
/* Maximum length of a string = 255 characters.*/  
/* When you need a small string of 15 characters. */  
/* Waiting for Trigger (WTG) = bit 5 = value 32 */  
int slot0;  
/* Device number of module in slot 0. Slot0 is configured in GPIB.Com as  
GPIB address 5, secondary address 96. */  
main( )  
{
char *cmd;  
char cmd_buff[MAX_LEN];  
char vpoint[SMALL_STRING];  
char vlist[MAX_LEN];  
char condition_data[SMALL_STRING];  
int i;  
/* Used to hold command strings sent to the module. */  
/* Used to hold command strings during string manipulations. */  
/* Used to hold the string equivalent of one voltage ramp step. */  
/* Used to hold the entire voltage List command header and points. */  
/* Reserve space for reading back status conditions. */  
/* Loop counter. */  
float vstart = 2.0;  
/* Start voltage for the ramp. */  
float vstop = 10.0;  
/* Stop voltage for the ramp. */  
float ramptime = 0.5;  
float dwell;  
/* Transition time for the ramp. */  
/* Dwell time for each ramp step. */  
dwell = ramptime / 19.0;  
/* Since the output stays at the last voltage point after its  
dwell expires, the dwell time of the last point is not part of the  
transition time. Therefore, divide the total time by 19 points, not 20.  
You want the same dwell time for every point in the List, so only  
download 1 dwell time. */  
if ((slot0 = ibfind(“SLOT0”)) < 0)  
finderr( );  
/* Assign unique identifier to the device slot0 and store in */  
/* variable slot0. Error = negative value returned.  
*/  
cmd = "*RST;*CLS;STATUS:PRESET”;  
ibwrt(slot0, cmd, strlen(cmd));  
if (ibsta & ERR)  
/* Reset and clear module. */  
error(cmd);  
sprintf(cmd_buff , "VOLT %f”, vstart);  
/*  
/* Start ramp at vstart. Use number to string conversion to send */  
real numbers over the bus as part of the command string.  
*/  
ibwrt(slot0, cmd_buff, strlen(cmd_buff));  
if (ibsta & ERR)  
error(cmd_buff);  
cmd = "CURR .1”;  
ibwrt(slot0, cmd, strlen(cmd));  
if (ibsta & ERR)  
error(cmd);  
cmd = "OUTPUT ON";  
ibwrt(slot0, cmd, strlen(cmd));  
if (ibsta & ERR)  
/* Enable output */  
error(cmd);  
Application Programs 107  
Download from Www.Somanuals.com. All Manuals Search And Download.  
cmd = "VOLT:MODE LIST”;  
ibwrt(sloto, cmd, strten(cmd));  
if (ibsta & ERR)  
/* Set to get voltage from List */  
error(cmd);  
strcpy(vlist, "LIST:VOLT “);  
/* Start with the command header for the voltage List. */  
/* The Loop calculates the string  
for (i = 1; i < NUM_PTS; i++) {  
*/  
*/  
sprintf(vpoint, “ %f, “, vstart+(((vstop - vstart) / NUM_PTS) * i));  
strcat(vlist, vpoint);  
/* equivalents of the voltage List  
/* points and concatenates them for */  
}
/* only the first 19 because there  
/* should not be comma after the  
/* last point. Do the last point  
/* separately with no comma.  
*/  
*/  
*/  
*/  
sprintf(vpoint , “%f” , vstop);  
strcat(vlist, vpoint);  
ibwrt(slot0, vlist, strlen(vlist));  
if (ibsta & ERR)  
/* Download voltage List points */  
error(vlist);  
sprintf(cmd_buff, "LIST:DWELL %f”, dwell);  
ibwrt(slot0, cmd_buff, strlen(cmd_buff));  
if (ibsta & ERR)  
/* Download 1 dwell time. Use number to */  
/* string conversion to send the real  
/* number over the bus as part of the  
/* command string.  
*/  
*/  
*/  
error(cmd_buff);  
cmd = "LIST:STEP AUTO”;  
ibwrt(slot0, cmd, strlen(cmd));  
if (ibsta & ERR)  
/* Dwelt-paced List */  
error(cmd);  
cmd = "INITIATE";  
ibwrt(slot0, cmd, strlen(cmd));  
if (ibsta & ERR)  
/* Enable trigger to start List */  
error(cmd);  
/* Before triggering the module, determine if it is ready by checking for  
'Waiting for Trigger' (bit 5 of the Operation Status Register).  
You could eliminate this step by simply inserting a pause in the program. However, by checking the instrument status, you can avoid  
timing problems. Also, any other operations that take time will give the module a chance to complete processing. */  
do  
{
cmd = "STATUS:OPERATION:CONDITION?";  
ibwrt(slot0, cmd, strlen(cmd));  
if (ibsta & ERR)  
error(cmd);  
ibrd(slot0, condition_data, SMALL_STRING);  
/* Allow to read SMALL_STRING bytes, which is more */  
if (ibsta & ERR)  
/* than enough. Note that first byte will be a + sign, */  
/* so you must convert the string to float, then to int, */  
error(condition_data);  
/* to do an integer bit test.  
*/  
} while (((int)(atof(condition_data)) && WTG) == 0); / * Loop until WTG = bit 5 (value 32) is true.  
/* Send trigger command to start List and generate the voltage ramp. */  
*/  
cmd = "TRIGGER:IMMEDIATE”;  
ibwrt(slot0, cmd, strlen(cmd));  
if (ibsta & ERR)  
/* This is an immediate trigger, which is always  
/* active. Therefore, it does not need to be  
/* selected as a trigger source.  
*/  
*/  
*/  
error(cmd);  
108 Application Programs  
Download from Www.Somanuals.com. All Manuals Search And Download.  
}
firiderr( )  
{
/* Indicates that ibfind failed */  
printf(“lbfind error: Does device name given match configuration name?\n”);  
}
error(bad_string)  
char *bad_string;  
/* This is a generalized error checking routine. */  
{
printf("GPIB error while sending or receiving ‘%s’.\n”, bad_string);  
printf("GPIB status: ibsta = 0x%x, iberr = 0x%x, ibcnt = 0x%x\n”, ibsta, iberr, ibcnt);  
}
Application Programs 109  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Index  
A
<AARD> .....................................................................................................................................................................15  
ANSI/IEEE, .......................................................................................................................................7-8, 23, 51, 54, 64  
C
CAL bit, ................................................................................................................................................................43, 52  
calibration password, ..................................................................................................................................................33  
CC bit, .......................................................................................................................................................43, 50, 52, 56  
commands  
common, ................................................................................................................................................10, 14, 23, 24  
completion, ..............................................................................................................................................................64  
diagram (see tree diagram)  
overlapped, ........................................................................................................................................................26, 64  
related, .....................................................................................................................................................................23  
CME bit, ...............................................................................................................................................................25, 52  
<CRD>, ......................................................................................................................................................................15  
CV bit, .......................................................................................................................................................43, 50, 52, 55  
D
data  
boolean, ...................................................................................................................................................................15  
character, ..................................................................................................................................................................15  
multiplier, ................................................................................................................................................................15  
numerical, ................................................................................................................................................................14  
suffix,........................................................................................................................................................................15  
DCL command,......................................................................................................................................................27, 31  
DDE bit, ................................................................................................................................................................25, 52  
default state (see *RST state)  
DFI output, ...........................................................................................................................................................49, 62  
DWE bit,..........................................................................................................................................................52, 56, 59  
DOS driver, .................................................................................................................................................................16  
E
error queue, ...........................................................................................................................................................24, 46  
error messages  
runtime,........................................................................................................................................................................65  
selftest, ..................................................................................................................................................................31, 65  
system, ........................................................................................................................................................................65  
ESB bit, ..............................................................................................................................................25, 30, 50, 52, 55  
event handle, .........................................................................................................................................................49, 54  
EXE bit, ................................................................................................................................................................25, 52  
F
factory-default state (see *RST state)  
FLT output (see DFI)  
Index 111  
Download from Www.Somanuals.com. All Manuals Search And Download.  
G
<GET>, .................................................................................................................................................................31, 47  
GWBASIC, ...........................................................................................................................................................17, 20  
H
header, ........................................................................................................................................................................11  
conventions, .........................................................................................................................................................12  
long form, ............................................................................................................................................................11  
optional, .........................................................................................................................................................12, 13  
path, .....................................................................................................................................................................13  
separator, .............................................................................................................................................................12  
short form, ...........................................................................................................................................................11  
GPIB address  
primary , ..........................................................................................................................................9, 11, 17, 18  
secondary, .............................................................................................................................................9, 17, 18  
setting, ...............................................................................................................................................................9  
Agilent BASIC, .......................................................................................................................................17, 18, 19  
GPIB capabilities, ..................................................................................................................................................9  
I
IEEE (see ANSI/IEEE)  
INH input (see RI)  
implied ABORT, ............................................................................................................................................28, 37, 60  
interrupt, source of, .....................................................................................................................................................54  
K
keyboard operation (see User’s Guide)  
keyword (see header)  
L
language, SCPI (see SCPI)  
link parameters, ...........................................................................................................................................................50  
local lockout command, ................................................................................................................................................7  
local sense switch, ......................................................................................................................................................48  
list, ..............................................................................................................................................................................61  
commands, .......................................................................................................................................35, 37, 38, 39, 48  
dwell-paced, .............................................................................................................................................................62  
intervals, ..................................................................................................................................................................61  
programming, ...........................................................................................................................................................61  
sequencing, ..............................................................................................................................................................61  
trigger-paced, ...........................................................................................................................................................62  
LSC pulse, ............................................................................................................................................................49, 59  
M
manuals, Series 66lxxA, ...............................................................................................................................................7  
MAV bit, ...................................................................................................................................................30, 50, 52, 54  
message terminator, ....................................................................................................................................................12  
message unit, ...............................................................................................................................................................10  
112 Index  
Download from Www.Somanuals.com. All Manuals Search And Download.  
message unit separator, ...............................................................................................................................................12  
message units, combining, ..........................................................................................................................................10  
Microsoft C, ..........................................................................................................................................................18, 19  
module identification, .................................................................................................................................................26  
module options, ..........................................................................................................................................................27  
MSS bit, ....................................................................................................................................................29, 30, 52, 54  
N
National Instruments DOS driver, ..............................................................................................................................18  
non-SCPI commands, .................................................................................................................................................67  
nonvolatile memory, .................................................................................................................................26, 28, 29, 30  
<NRF>,........................................................................................................................................................................15  
<NRF+>,......................................................................................................................................................................15  
O
OC bit, ......................................................................................................................................................36, 41, 45, 56  
OCP, .....................................................................................................................................................................36, 41  
OPC bit, ..................................................................................................................................25, 26, 41, 52, 63, 72, 74  
OPER bit, ......................................................................................................................................30, 45, 50, 52, 55, 56  
OT bit, ......................................................................................................................................................41, 45, 52, 56  
output queue, ........................................................................................................................................................27, 54  
overcurrent protection (see OCP)  
overlapped commands, .........................................................................................................................................26, 63  
overvoltage protection (see OVP)  
OV bit,.............................................................................................................................................................45, 50, 52  
P
parallel commands (see commands, overlapped)  
parameters, ................................................................................................................................................23, 24, 31, 50  
passcode, calibration (see calibration password)  
pending operations, ...............................................................................................................................................27, 31  
PON bit, ..........................................................................................................................................................25, 52, 55  
power-on status (see *RST state)  
primary address (see GPIB address)  
program message, .......................................................................................................................................................10  
programming voltage, ...........................................................................................................................................77, 78  
PSC , ...............................................................................................................................................................25, 27, 54  
Q
query, ....................................................................................................................................................................12, 23  
QUES bit, .................................................................................................................................................30, 45, 50, 52  
QYE bit, ................................................................................................................................................................25, 52  
R
reading registers,..........................................................................................................................................................55  
recalled parameters, ....................................................................................................................................................28  
reference documents, ....................................................................................................................................................7  
remote inhibit (see RI)  
remote sense switch, ...................................................................................................................................................49  
Index 113  
Download from Www.Somanuals.com. All Manuals Search And Download.  
reset, parameters, ........................................................................................................................................................50  
reset state (see *RST state)  
response message, .......................................................................................................................................................10  
RI  
configuration switch for (see Chapter 2 in User’s Guide)  
description of,.......................................................................................................................................................62  
digital connector pins for (see Chapter 3 in Installation Guide)  
examples of use (see Chapter 4 in User’s Guide)  
example of wiring (see Chapter 3 in Installation Guide)  
RI Questionable Status bit, ......................................................................................................................41, 45, 52  
signal electrical characteristics (see Chapter 1 in Installation Guide)  
root specifier, ..............................................................................................................................................................12  
RQS bit, ..............................................................................................................................................29, 30, 52, 53, 55  
*RST state, ............................................................................................................................................................28, 50  
RTG pulse, ......................................................................................................................................................50, 60, 63  
S
saved parameters, ........................................................................................................................................................29  
SCPI, .............................................................................................................................................................................9  
confirmed commands, .............................................................................................................................................67  
version, .............................................................................................................................................................49, 67  
secondary address (see GPIB address)  
sequential commands, .................................................................................................................................................63  
serial poll, ...................................................................................................................................................................4-3  
SETUP.BAS, .............................................................................................................................................................2-8  
SRQ, .........................................................................................................................................................27, 30, 54, 55  
STC pulse, ......................................................................................................................................................43, 50, 60  
STS pulse,....................................................................................................................................................... 49, 60, 63  
SUM3 bit, ...................................................................................................................................................................49  
T
TDC pulse, ......................................................................................................................................................49, 60, 63  
tree diagram, ..........................................................................................................................................................31-32  
Trigger In, ...................................................................................................................................................................60  
trigger  
delay, .....................................................................................................................................................................58  
fixed mode, ............................................................................................................................................................58  
initiating, ...............................................................................................................................................................58  
list mode, .........................................................................................................................................................58, 59  
Trigger Out, ................................................................................................................................................................60  
fixed, .....................................................................................................................................................................58  
list, .........................................................................................................................................................................59  
U
units (see data suffix)  
UNR bit, ...............................................................................................................................................................45, 52  
W
WTG bit, .....................................................................................................................................32, 43, 47, 52, 59, 105  
114 Index  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Agilent Sales and Support Offices  
For more information about Agilent Technologies test and measurement products, applications, services,  
and for a current sales office listing, visit our web site: http://www.agilent.com/find/tmdir  
You can also contact one of the following centers and ask for a test and measurement sales  
representative.  
United States:  
Latin America:  
Agilent Technologies  
Test and Measurement Call Center  
P.O. Box 4026  
Englewood, CO 80155-4026  
(tel) 1 800 452 4844  
Agilent Technologies  
Latin American Region Headquarters  
5200 Blue Lagoon Drive, Suite #950  
Miami, Florida 33126  
U.S.A.  
(tel) (305) 267 4245  
(fax) (305) 267 4286  
Canada:  
Australia/New Zealand:  
Agilent Technologies Canada Inc.  
5150 Spectrum Way  
Mississauga, Ontario  
L4W 5G1  
Agilent Technologies Australia Pty Ltd  
347 Burwood Highway  
Forest Hill, Victoria 3131  
(tel) 1-800 629 485 (Australia)  
(fax) (61 3) 9272 0749  
(tel) 1 877 894 4414  
(tel) 0 800 738 378 (New Zealand)  
(fax) (64 4) 802 6881  
Europe:  
Asia Pacific:  
Agilent Technologies  
Test & Measurement European Marketing Organisation  
P.O. Box 999  
1180 AZ Amstelveen  
The Netherlands  
Agilent Technologies  
24/F, Cityplaza One, 1111 King’s Road,  
Taikoo Shing, Hong Kong  
tel: (852)-3197-7777  
fax: (852)-2506-9284  
(tel) (31 20) 547 9999  
Japan:  
Agilent Technologies Japan Ltd.  
Measurement Assistance Center  
9-1, Takakura-Cho, Hachioji-Shi,  
Tokyo 192-8510, Japan  
(tel) (81) 426 56 7832  
(fax) (81) 426 56 7840  
Technical data is subject to change.  
115  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Manual Updates  
The following updates have been made to this manual since the print revision indicated on the title page.  
4/15/00  
All references to HP have been changed to Agilent.  
All references to HP-IB have been changed to GPIB.  
Information about VXIPlug&Play instrument drivers has been added to chapter 1.  
Download from Www.Somanuals.com. All Manuals Search And Download.  

ADTRAN Network Router 1200130L2 User Manual
Air King Humidifier AK4500 User Manual
Airlink101 Switch AGIGA5SW User Manual
Airlink Switch UG AGIGA5SW 1105 User Manual
Allied Telesis Switch AT 8624POE User Manual
Asus Computer Accessories 1 E1309 User Manual
Audiovox Satellite Radio XAPH1 User Manual
Axis Communications Printer AXIS 5400+ User Manual
Baldor Riding Toy mn722 User Manual
Behringer Speaker BEHRINGERPSU SB User Manual