Agilent Technologies Graphics Tablet DC122 User Manual

XZ060  
AGILENT ACQIRIS  
USER MANUAL  
FAMILY OF  
10-BIT DIGITIZERS  
For Use with DC Series Digitizers  
Models covered:  
DC122 / DC152  
DC222 / DC252 / DC282  
User Manual: Family of 10-bit Digitizers  
ZM020170B Rev.D  
Download from Www.Somanuals.com. All Manuals Search And Download.  
CONTENTS  
User Manual: Family of 10-bit Digitizers  
Page 3 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3.7. ASBus2 ....................................................................................................................................... 34  
User Manual: Family of 10-bit Digitizers  
Page 4 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1.  
Out of the Box  
1.1.  
Message to the User  
Congratulations on having purchased an Agilent Technologies Acqiris data conversion product. Acqiris Digitizers  
are high-speed data acquisition modules designed for capturing high frequency electronic signals. To get the most out  
of the products we recommend that you read this manual carefully. We trust the product you have purchased will  
meet with your expectations and provide you with a high quality solution to your data conversion applications.  
1.2.  
Using this Manual  
This guide assumes you are familiar with the operation of a personal computer (PC) running a Windows 2000/XP or  
other supported operating system. It also assumes you have a basic understanding of the principles of data acquisition  
using either a waveform digitizer or a digital oscilloscope.  
The manual is divided into 5 separate sections. To understand the elements of operation for the module it is essential  
that you read them all.  
Chapter 1  
Chapter 2  
OUT OF THE BOX, describes what to do when you first receive your new Acqiris product.  
Special attention should be paid to sections on safety, packaging and product handling. Before  
installing your product please ensure that your system configuration matches or exceeds the  
requirements specified.  
INSTALLATION, covers all elements of installation and performance verification. Before  
attempting to use your Acqiris product for actual measurements we strongly recommend that you  
read all sections of this chapter.  
Chapter 3  
Chapter 4  
PRODUCT DESCRIPTION, provides a full description of all the functional elements of the  
digitizer.  
RUNNING THE ACQIRISLIVE APPLICATION, describes the operation of AcqirisLive 3.1, an  
application that enables basic operation of Acqiris digitizers in a Windows 2000/XP environment.  
Note: AcqirisMAQS is an alternate software application offering many interesting possibilities for  
the control of acquisition systems in a single or multi-machine environment. Ask your salesman or  
Acqiris for more information.  
Chapter 5  
RUNNING THE GEOMAPPER APPLICATION, describes the purpose and operation of the  
GeoMapper application which is needed for some ASBus2 Multi-instrument systems.  
For information necessary for writing you own software to control Acqiris products you should refer to the  
Programmer’s Guide and the Programmer’s Reference Manual.  
1.3.  
Conventions Used in This Manual  
The following conventions are used in this manual:  
This icon to the left of text warns that an important point must be observed.  
WARNING  
CAUTION  
Denotes a warning, which advises you of precautions to take to avoid being electrically shocked.  
Denotes a caution, which advises you of precautions to take to avoid electrical, mechanical, or  
operational damages.  
NOTE  
Italic  
Denotes a note, which alerts you to important information.  
text denotes a warning, caution, or note.  
Bold Italic  
mono  
text is used to emphasize an important point in the text or a note  
text is used for sections of code, programming examples and operating system commands.  
1.4.  
Disclaimer and Safety  
The 10-bit DC Series CompactPCI/PXI Digitizers have been designed to operate inside a CompactPCI/PXI crate.  
The crate provides the modules with all needed power. Agilent Technologies does not recommend operation of the  
DC Series modules outside of a CompactPCI/PXI crate.  
User Manual: Family of 10-bit Digitizers  
Page 5 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
CAUTION:  
Do not exceed the maximum input voltage rating! The maximum input voltage for 50 Ω input  
impedance is ±5 V. The maximum input for 1 MΩ input impedance is ±300 V (dc + ac).  
1.5.  
Warning Regarding Medical Use  
The Digitizer cards are not designed with components and testing procedures that would ensure a level of reliability  
suitable for use in treatment and diagnosis of humans. Applications of these cards involving medical or clinical  
treatment can create a potential for accidental injury caused by product failure, or by errors on the part of the user.  
These cards are not intended to be a substitute for any form of established process or equipment used to monitor or  
safeguard human health and safety in medical treatment.  
WARNING:  
The modules discussed in this manual have not been designed for making direct measurements  
on the human body. Users who connect an Acqiris module to a human body do so at their own  
risk.  
1.6.  
Packaging and Handling  
Your Digitizer is shipped with the following components:  
A small 3.5” compact disc in an Acqiris paper CD envelope that includes  
o
10 product user manuals in electronic form (Signal Analyzers, Streamer Analyzers, Family of 8-bit  
Digitizers, Family of 10-bit Digitizers, Family of 12-bit Digitizers, Family of Averagers, Family of  
Analyzers, Family of Time Counters, CC10X Family of CompactPCI Crates and CC121  
CompactPCI Crate),  
o
o
o
o
o
o
o
o
o
1 Programmer’s Guide and 1 Programmer’s Reference Manual,  
device drivers with sample software for different operating systems, environments and languages,  
the AcqirisAnalyzers application, a demonstration program for the AC/SC Analyzer products,  
the AcqirisLive application, a demonstration program for our digitizer and averager products,  
the DemoSSR application, a demonstration program for the Acqiris AP235/AP240 Analyzers,  
the DemoAPX01 application, a demonstration program for the Acqiris AP101/AP201 Analyzers,  
the DemoTC application, a demonstration program for the Acqiris TC840/TC890 Time Counters,  
product data sheets,  
full installation procedures for use with Microsoft Windows, National Instruments LabVIEW RT,  
Wind River VxWorks, IVI-COM/C, and Linux software.  
A declaration of conformity  
Optional documentation such as a model-dependent document giving Specifications & Characteristics, a  
Calibration Certificate, or a Performance Verification  
After carefully unpacking all items, inspect each to ensure there are no signs of visible damage. Also check that all  
the components received match those listed on the enclosed packing list. Agilent Technologies cannot accept  
responsibility for missing items unless we are notified promptly of any discrepancies. If any items are found to be  
missing or are received in a damaged condition please contact the Agilent service center or your local supplier  
immediately. Retain the box and packing materials for possible inspection and/or reshipment.  
1.7.  
Warranty  
All Agilent Technologies Acqiris Digitizer products are warranted to operate within specification, assuming normal  
operation, for a period of three years from the date of shipment. It is recommended that yearly calibration be made in  
order to verify product performance. All repairs, replacement and spare parts are warranted for a period of 3 months.  
A 5-year repair warranty is available as an option.  
Agilent Technologies endeavors to provide leading edge technology that includes the latest concepts in hardware and  
software design. As such software and firmware used with the products is under continual refinement and  
improvement. All software and instrument firmware is supplied “as is” with no warranty of any kind. Software and  
firmware is thoroughly tested and thought to be functional at the time of shipment. At Agilent Technologies’  
discretion software and firmware may be revised if a significant operational malfunction is detected.  
Products supplied but not manufactured by Agilent Technologies are covered solely by the warranty of the original  
equipment manufacturer.  
User Manual: Family of 10-bit Digitizers  
Page 6 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
In exercising this warranty, Agilent Technologies will repair or replace any product returned to the Agilent service  
center, within the warranty period. The warranty covers all defects that are a result of workmanship or materials. This  
excludes defects that are caused by accident, misuse, neglect, or abnormal operation.  
The purchaser is responsible for returning the goods to the nearest Agilent service center. This includes  
transportation costs and insurance. Agilent Technologies will return all warranty repairs with transportation prepaid.  
1.8.  
Warranty and Repair Return Procedure, Assistance and Support  
Agilent Technologies Inc. acquired Acqiris SA and its product lines in December 2006. Before returning any Agilent  
Technologies Acqiris product for repair please contact your nearest Agilent Technologies Service Center.  
You can find information about technical and professional services, product support, and equipment repair and  
service on the Web, see http://service.tm.agilent.com (or http://www.agilent.com/ and after selecting your country  
click on Contact Us). The service center will ask for your name, company, phone number and address, the model  
and serial numbers of the unit to be repaired, and a brief description of the problem.  
Before issuing a Service Order they will ask you to communicate with us by phone or eMail so that we can learn as  
much as needed about the problems observed. If a unit returned under guarantee is found to be working normally and  
this procedure was not followed we reserve the right to charge you for the work done.  
For your nearest customer support center contact Acqiris at 1-877-ACQIRIS in the USA, +41 22 884 32 90 in  
Europe or +61 3 9210 2890 in the Asia-Pacific region. Alternatively, please contact Acqiris Technical Support  
(support@acqiris.com) or come visit our web site at http://www.acqiris.com. The Agilent Technologies Support  
Centers can also help redirect you for any questions concerning the installation and operation of your equipment.  
1.9.  
System Requirements  
In order to obtain reasonable performance from your digitizer, Acqiris products need the following minimum PC  
System Requirements:  
Processor: 150 MHz Pentium (higher recommended)  
Memory: 64 Mbytes RAM. The previous number is a very rough estimate. Assuming that you are using AcqirisLive  
or an application of your own that operates on the acquired data it seems reasonable to ask for 10 times the  
total acquisition memory that you will be using at the same time in the application. Performance is likely  
to be degraded if less memory is available.  
Display resolution: At least 800 x 600 pixels and 256 colors for use of AcqirisLive or AcqirisDemo  
Operating System: Microsoft Windows 2000/XP including 2003 Server, Wind River VxWorks 5.5.1, and Linux.  
The following Linux versions have compiled loadable kernel modules:  
ƒ
ƒ
ƒ
ƒ
ƒ
ƒ
RedHat Enterprise Linux Version 3 - Kernel Version 2.4.21-4.EL  
RedHat Enterprise Linux Version 3 - Kernel Version 2.4.21-4.ELsmp,  
Suse Linux 9.2 - Kernel version 2.6.8-24-smp,  
Debian Sarge 3.1 2.6.8-3-686,  
Debian 4.0 etch Kernel version 2.6.18-4-686,  
Scientific Linux 4.4 Kernel version 2.6.9-11.EL).  
Linux kernel driver source code is available for compilation. Support for Windows 95/98/NT4 is included “as is”  
since these operating systems are no longer supported by Microsoft.  
Hard Drive Space: 20 Mbytes Minimum  
CD Drive (or any method to copy Acqiris Software installation files from CD to the hard drive such as LAN, floppy  
drive, etc.)  
LabVIEW: The Acqiris LabVIEW driver is available for National Instruments LabVIEW versions 7.1or 8.0  
LabVIEW RT: The Acqiris LabVIEW RT driver is available for National Instruments LabVIEW RT version 7.1 or  
higher. The VISA driver must be version 3.0 or higher (and not 3.2,3.3, or 3.3.1 for DC2x2).  
MATLAB: The Acqiris MEX interface can be used with MathWorks MATLAB 6.5 or a newer version.  
Visual BASIC: The interface files and examples are available for Microsoft Visual Basic versions 5 or 6 and the  
interface files only for .NET.  
Tornado: The example files are useable with Wind River Tornado 2.2.1  
User Manual: Family of 10-bit Digitizers  
Page 7 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
1.10. Transport & Shipping  
CAUTION: Cards can be safely transported in their original shipping packages. DC cards can be transported  
when properly mounted in a CompactPCI crate.  
To package the instrument for shipping:  
Step  
Notes  
1. Place the instrument in its original packaging If the original packaging materials are not available,  
materials.  
use a professional packaging service. Contact your  
Agilent Service Center for more information.  
2. Surround the instrument with at least 3 to 4 inches (8  
to 10 cm) of its original packing material or bubble-pack  
to prevent the instrument from moving in its shipping  
container.  
3. After wrapping it with packing material, place the The shipping container must be large and strong  
instrument in its original shipping container or a strong enough to accommodate your instrument and allow at  
shipping container that is made of double-walled least 3 to 4 inches (8 to 10 cm) on all sides for packing  
corrugated cardboard with 159 kg (350 lb) bursting material.  
strength.  
4. Seal the shipping container securely with strong nylon  
adhesive tape.  
5. Mark the shipping container “FRAGILE, HANDLE  
WITH CARE” to help ensure careful handling.  
6. Use the address obtained from your Agilent  
Technologies Service Center.  
7. Retain copies of all shipping papers.  
CAUTION: Damage can result if the original packaging materials are not used. Packaging materials should be  
anti-static and cushion the instrument on all sides. NEVER USE STYRENE PELLETS IN ANY SHAPE AS  
PACKAGING MATERIALS. They do not adequately cushion the instrument or prevent it from moving in the  
shipping container. Styrene pellets can also cause equipment damage by generating static electricity or by lodging  
in fan motors.  
1.11. Maintenance  
The cards do not require any maintenance. There are no user serviceable parts inside. A periodic calibration can be  
obtained on request.  
1.12. Cleaning  
Cleaning procedures consist only of exterior cleaning.  
Clean the exterior surfaces of the module with a dry lint-free cloth or a soft-bristle brush. If any dirt remains, wipe  
with a cloth moistened in a mild soap solution. Remove any soap residue by wiping with a cloth moistened with clear  
water. Do not use abrasive compounds on any parts.  
1.13. Disposal and Recycling  
Electronic equipment should be properly disposed of. Acqiris Digitizers and their accessories must not be thrown out  
as normal waste. Separate collection is appropriate and may be required by law.  
User Manual: Family of 10-bit Digitizers  
Page 8 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
2.  
Installation  
This chapter describes how to install the Acqiris hardware and software for Windows 2000/XP, Linux, National  
Instruments LabVIEW RT, Linux, or Wind River VxWorks.  
NOTE: For a first time installation we strongly recommend installing the software before inserting the  
hardware into the PC.  
2.1.  
IC414 Installation  
NOTE: If you are going to install an IC414 interface for the first time and are running Windows 2000 or XP you  
should follow the procedure below before installing the Acqiris hardware.  
2.1.1. IC414 Hardware installation hints  
The PCI-8570/PXI-8570 User's Manual (Rev. 1.00) section 2.5 gives Hardware Installation instructions.  
CAUTION: Turn off the power of the PC; the PC may have to be unplugged to ensure that the PCI bus has no  
power available.  
CAUTION: Touch the antistatic package to a grounded object before removing the card from the package.  
Electrostatic discharge can damage the card.  
The standard cable pair provided each have a red connector on one end and a black connector on the other. Therefore  
the correct connection can be made by plugging the Red connector into the L0Rx socket and the Black connector into  
the L0Tx socket on the PXI module and the other Red connector into the PCI module socket furthest from the PCI  
card internal base connector and the Black connector into the next socket.  
If you intend to use 64-bit 66 MHz transfer to maximize data transfer speed you should cable a “bundled link”, using  
two standard cable pairs and both the L0 and L1 pairs of connectors. You should also make sure that you configure  
the PXI-8570 M66EN Jumper correctly.  
2.1.2. IC414 Windows software installation  
Linux users do not need to read any further since there is no special software installation.  
Windows users should have the hardware installed as noted above. This software installation should be done before  
any Acqiris modules are placed in the CompactPCI crates. This may mean that you have to remove the module from  
the crate as delivered.  
The crate should be turned on first followed by the PC. If the cabling and start-up sequence is done correctly there  
will be no LED illuminated on the PCI unit connected pair and the LED's of the PXI connected pair will be lit.  
For Windows XP installation, Select the Control Panel under Settings in the Start menu. Then, if you are using the  
Category View select Printers and Other Hardware. After this, for both Category and Classic views, go to  
System and then display the Hardware tab to get access to the Driver Signing menu. Since neither the AdLink nor  
the Acqiris driver has been submitted for Windows Logo testing you must select either the Ignore or Warn action.  
The resulting menu looks as shown:  
User Manual: Family of 10-bit Digitizers  
Page 9 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
The PCI-8570/PXI-8570 User's Manual (Rev. 1.00) section 2.4 contains the software installation instructions. These  
should be executed before allowing the hardware installation process to look for the driver. If you have an AdLink  
CD Version 2004A4 or later you can use it; if not you should download the latest driver from the WEB site  
(http://www.adlinktech.com/). You can then continue with the Hardware Installation. A reboot will then be  
necessary. At this point the Stargen Fabric PCI Adapter and the Stargen Aruba Fabric-to-PCI Bridge should appear  
correctly installed under System Devices in the Device Manager.  
NOTE: If you have an AdLink CD Version 2005A3 or later you can find 8570install.exe in the folder X:\Driver  
Installation\PXI Platform\PXI Extension\PCI_PXI-8570\Wnt2kxp and the starfab1.inf file in the folder X:\Driver  
Installation\PXI Platform\PXI Extension\PCI_PXI-8570\Win98.  
2.2.  
Installing the Software under Windows  
2.2.1. Warnings  
If Setup detects a previous installation of Acqiris software on your system, a warning screen will be displayed. It is  
recommended to exit Setup and uninstall older versions.  
The installer from software releases prior to Acqiris Software 2.0 installed the Digitizer Driver DLL files into the  
System directory. These will be removed by Setup. If you wish to keep the old installation on your system, you  
should exit Setup, and move all Acqiris driver files (acqiris*, acqrs* and acqir*) to some archive directory.  
The DLL files will be installed into the bin subdirectory of the Acqiris software root, and the corresponding path will  
be added to the PATH environment variable.  
2.2.2. Multiple Versions  
With the software installation from Acqiris Software 2.0 (or above), it is possible to keep multiple versions on the  
same system, but you must specify a different root directory (i.e. Install Folder). If you keep the same directory,  
Setup will overwrite your previous installation.  
To go back to a previous version, you must change the PATH environment variable and reinstall the Kernel driver.  
Note that the 10-bit instruments are not supported by Kernel versions before 1.8 which was part of Acqiris Software  
2.13c. Under Windows 2000/XP,  
1. Copy the SYS file from <old_AcqirisSoftware_root>\bin\kernelto the  
Windows\System32\driversdirectory.  
2. Change the AcqirisDxRoot, AcqirisDxDir and PATH environment variables to the old root.  
3. Reboot the computer.  
User Manual: Family of 10-bit Digitizers  
Page 10 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
2.2.3. Installation  
Before installing the Acqiris hardware, you should complete the following steps to install the software for Windows  
2000/XP.  
NOTE: You will need administrator privileges to complete the software installation under Windows  
2000//XP.  
1. Insert the Acqiris Software CD into the CD-ROM drive of your computer. If the Autorun program does not start  
automatically (Autoplay disabled), you can start it manually, or navigate to the AcqirisSoftware\Windows folder  
in order to display the files included.  
2. Choose Install AcqirisSoftware for Windows 2000/XP (or run Setup.exe from the  
AcqirisSoftware\Windows folder). After several seconds for initialization the following screen will appear.  
User Manual: Family of 10-bit Digitizers  
Page 11 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
3. To proceed with the installation click “Next”. The following screen will be shown.  
4. Please enter your user information and click “Next” to continue. If the program finds that there is still Acqiris  
software installed on your machine a warning panel (not shown) will appear. Otherwise, the next screen shows:  
5. Pick the desired combination of module families and click "Next" to continue.  
User Manual: Family of 10-bit Digitizers  
Page 12 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
6. The screen above will normally allow the documentation to be installed. Remove the check from the box if you  
do not want online access to the manuals.  
7. The next screen allows you to enable LabVIEW RT and/or Wind River VxWorks support. By default there will  
be none but if desired you can install any of them together with Windows support or without Windows support.  
Click “Next” to continue.  
User Manual: Family of 10-bit Digitizers  
Page 13 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8. The screen below asks for the desired installation type. After having made your choice Click “Next” to continue.  
9. If you chose the Custom installation, the following screen will let you select each package individually. Note  
that the space indicated for LabVIEW, Firmware and UserManual packages is incorrect. The correct values are  
4.1 MB, 21 MB, and 27 MB respectively.  
10. If MATLAB is installed on your machine, you will be asked to point the installer to the MATLAB root  
directory. You should do this if you want the installer to modify the standard startup.m file to incorporate the  
Acqiris adaptor.  
11. Now that the elements of the installation have all been decided you will be prompted for the installation folder.  
This will be the root directory of the Acqiris software installation. If User Manuals (27 MB) and Firmware (21  
MB) are loaded more space than indicated here will be required on the drive. For the case of a Tornado 2.2  
installation the folder name should not contain any spaces.  
User Manual: Family of 10-bit Digitizers  
Page 14 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
12. Furthermore, you should give a name to the shortcut folder. This is the menu entry under Start Programs  
where you will find the shortcuts for AcqirisLive, manual(s), etc.  
13. If you have enabled the installation of 12-bit Digitizers, Averagers, or Analyzers, the next screen lets you change  
the name of the directory where the FPGA firmware will be installed.  
User Manual: Family of 10-bit Digitizers  
Page 15 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
14. AcqirisLive needs the LabWindows/CVI 7.0 Run-Time Engine to run. If Setup has detected that a  
LabWindows/CVI Run-Time Engine is already installed on your system, it will ask you if you would like to  
install it locally for AcqirisLive anyway. If you are not sure about the version of the CVI Run-Time Engine on  
your system, it is recommended to install it locally. Click “Next” to continue.  
15. Depending on the install type, you may be asked which LabVIEW version format you want for the LabVIEW  
files. Select the appropriate format and click “Next” to continue.  
User Manual: Family of 10-bit Digitizers  
Page 16 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
16. A summary will be shown to allow you to check what you have asked for  
17. You are now ready to install. You may still go back to any previous screen to modify your selection. Click  
“Install” when ready.  
User Manual: Family of 10-bit Digitizers  
Page 17 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
18. Setup will now copy the files and make the necessary changes to your system. When done, an information  
screen will be displayed. Please read this carefully.  
19. Registration of your installation will help us provide you with better support. You will also be notified of  
updates and upgrades. All information submitted to us will be treated confidentially and never be disclosed  
outside the company.  
User Manual: Family of 10-bit Digitizers  
Page 18 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
20. Setup will prepare a registration e-mail in your e-mail client application upon termination of the setup procedure.  
You can then decide whether or not you wish to send it. You may also add comments. Uncheck the box if you  
do not want to register your installation.  
User Manual: Family of 10-bit Digitizers  
Page 19 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
21. Click “Finish”. The software installation is now complete.  
22. You can now either accept the suggestion to restart the computer or you should shutdown your computer and  
proceed with the hardware installation.  
2.3.  
Installing the Software for Linux  
The Acqiris Software is ready to install and run on Linux systems with any of the following:  
ƒ
ƒ
ƒ
RedHat Enterprise Linux Version 3 - Kernel Version 2.4.21-4.EL  
RedHat Enterprise Linux Version 3 - Kernel Version 2.4.21-4.ELsmp,  
Suse Linux 9.2 - Kernel version 2.6.8-24-smp,  
User Manual: Family of 10-bit Digitizers  
Page 20 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
ƒ
ƒ
ƒ
Debian Sarge 3.1 2.6.8-3-686,  
Debian 4.0 etch Kernel version 2.6.18-4-686,  
Scientific Linux 4.4 Kernel version 2.6.9-11.EL)  
The following tar files exist with the driver and library compiled with the GNU gcc version shown:  
ƒ
ƒ
ƒ
AcqirisLinux-3.1a-gcc-3.3.tar.gz - compiled under Suse 9.2 with GNU gcc 3.3.  
AcqirisLinux-3.1a-gcc-3.4.tar.gz - compiled under SL 4.4 with GNU gcc 3.4.  
AcqirisLinux-3.1a-gcc-4.1.tar.gz - compiled under Debian etch with GNU gcc 4.1.  
The Kernel Driver source code is available for recompilation for users with other Linux systems.  
The appropriate tar file should be copied to a local directory (e.g. your home directory) and then unpacked by using  
the following command  
tar xzf AcqirisLinux-3.1a-gcc-X.x.tar.gz  
The resulting directory AcqirisLinux contains an install script drv-install and a graphical Demo program  
demo/AcqirisDemo.  
Before installing the driver you have to put the correct version of the Kernel mode driver into acqirisPCI.o. Type  
cd AcqirisLinux/lib/modules  
and either  
cp acqrsPCI.o[.n] acqrsPCI.o  
(Linux kernel 2.4)  
(Linux kernel 2.6)  
or  
cp acqrsPCI.ko[.n] acqrsPCI.ko  
where [.n] is the running version of the Linux kernel. This can be obtained by running the command  
uname -r  
The following versions exist:  
acqrsPCI.o.2.4.21-4.EL  
acqrsPCI.ko.2.6.8-3-686  
acqrsPCI.ko.2.6.8-24-smp  
acqrsPCI.ko.2.6.9-11.EL  
acqrsPCI.ko.2.6.18-4-686  
for Red Hat Enterprise Linux version 3  
for Debian 3.1 'sarge'  
for Suse 9.2  
for Scientific Linux 4.4  
for Debian 4.0 'etch'  
To install the driver and the load script to the system, you have to get super user privileges and execute the driver  
install script by typing,  
cd AcqirisLinux  
./drv-install add  
You can check that the driver is loaded properly with lsmodor dmesg. The drv-install script has the following  
additional functionality:  
drv-install rem to remove the driver and load script  
drv-install res to restart the driver  
If you intend to use an Averager, an Analyzer, or a 12-Bit Digitizer you must install the Firmware .bit files. You can  
either copy them from the Firmware directory of the CD-ROM or download them from our WEB site. They should  
be placed in the Firmware subdirectory of AcqirisLinux. In the case of the CD-ROM, this can be done by continuing  
the above command sequence with the following:  
cp /mnt/CDRom/Firmware/*.bit Firmware/  
2.3.1. Kernel Mode Driver Compilation  
For some special kernels, for example SMP, the kernel mode driver will have to be recompiled on the target system.  
All files needed to compile a new kernel mode driver are in AcqirisLinux/linuxdriverpci for Kernel 2.4 or in  
AcqirisLinux/linux2.6driverpci for Kernel 2.6  
The makefile can be used for compilation under a standard compatible distribution. To compile the driver, the kernel  
header files need to be installed. Furthermore, depending on the Linux distribution, the kernel source files may need  
to be installed. The makefile will try to guess where they are located. If it fails, see instructions inside the makefile  
itself to specify manually where these kernel header files are located.  
To compile a new kernel under linux-2.4, issue the following commands:  
ƒ
cd linuxdriverpci  
User Manual: Family of 10-bit Digitizers  
Page 21 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
ƒ
ƒ
make clean to remove all *.o files in AcqirisLinux/linuxdriverpci.  
make to generate a new kernel mode driver acqrsPCI.o and also copy it to the directory  
AcqirisLinux/lib/modules where the install script (drv_install) can access it.  
ƒ
ƒ
drv_install rem to remove the previous installed kernel mode driver.  
drv_install add to install the new driver.  
To compile a new kernel under linux-2.6, issue the following commands:  
ƒ
ƒ
ƒ
cd linux2.6driverpci  
make clean all to generate a new kernel mode driver acqrsPCI.ko from scratch.  
make install to copy the kernel module where it should reside.  
Note that this version of the loadable kernel module had been tested on Linux kernel versions up to 2.6.18. However,  
starting with kernel 2.6.17, a few kernel functions relating to the device classes are no longer available to proprietary  
modules. Thus automatic creation of the /dev/acqrsPCI node has to be done manually in the /etc/rc.local file.  
For Debian users, in order to compile the kernel module on a system, a few packages that match the current kernel-  
image must be installed. For instance, on a Debian 3.1 sarge system 686 the required packages are:  
ƒ
ƒ
ƒ
ƒ
ƒ
ƒ
kernel-image-2.6-686  
kernel-image-2.6.8-3-686  
kernel-headers-2.6-686  
kernel-headers-2.6.8-3  
kernel-headers-2.6.8-3-686  
kernel-kbuild-2.6-3  
2.3.2. Special cases  
If you are running a Linux distribution that doesn’t use the standard paths for the load scripts, you can load the driver  
with the command  
cd AcqirisLinux  
./drv-install load  
Copy the driver to the module directory as follows:  
cp lib/modules/acqrsPCI.o[.n] /lib/modules/{kv}/ACQIRIS/acqrsPCI.o  
where [.n] is the optional designation of the desired version of the acqrsPCI driver as given above and {kv} is the  
appropriate system kernel version (obtained with uname -r, i.e. 2.4.20).  
Then add the following command to /etc/rc.local.  
/sbin/insmod –f /lib/modules//{kv}/ACQIRIS/acqrsPCI.o  
For systems using udev (the dynamic /dev/ hierarchy) it may be convenient to automatically create the /dev/acqrsPCI  
node at boot. To do this, add the following command to the /etc/rc.local file:  
/sbin/mknod -m 666 /dev/acqrsPCI c 124 0  
2.3.3. Environment variables for the Firmware  
Automatic loading of the firmware needed by 12-bit, analyzer, and averager modules relies on the environment  
variable AcqirisDxDir pointing to the directory containing the file AqDrv4.ini which in turn points to the directory  
containing the Firmware .bit files. Therefore, assuming that your Firmware is in /usr/local/AcqirisLinux/Firmware  
and that AqDrv4.ini is in /usr/local/AcqirisLinux/demo then you must edit AqDrv4.ini so that it contains the line  
fpgaPath=/usr/local/AcqirisLinux/Firmware  
Then, if your shell is csh or tcsh modify the /etc/csh.login file to contain the line  
setenv AcqirisDxDir /usr/local/AcqirisLinux/demo  
or, if your shell is bash, ksh, zsh or sh, modify the /etc/profile file to contain the lines  
AcqirisDxDir=/usr/local/AcqirisLinux/demo  
export AcqirisDxDir  
User Manual: Family of 10-bit Digitizers  
Page 22 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
2.4.  
Installing the Hardware  
1. Turn off the power of the PC and the crate in the case of a CompactPCI module. For PCI modules the PC  
should be unplugged to ensure that the PCI bus has no power available. However, CompactPCI crates can  
be left plugged in since this ensures proper grounding.  
CAUTION: Touch the antistatic package to a grounded metal part of the PC or crate before removing the  
card from the package. Electrostatic discharge can damage the card.  
2. Module in a CompactPCI crate: Follow the instructions of the crate manufacturer to insert the DC Series  
card into a free 6U CompactPCI peripheral slot. Be sure to ground yourself by touching the grounded crate  
and avoid touching any components on the DC Series card. Be sure to tighten both front panel mounting  
screws to lock the module into place and insure proper grounding of the frame.  
NOTE: To ensure the best possible performance, users of Acqiris CC121 Crates with ASBus2 systems should  
respect the module placement rules to be found in the Acqiris CC121 CompactPCI Crate User Manual.  
3. Turn on the power of the crate(s), and then the PC and start the operating system.  
NOTE: Acqiris Digitizers are equipped with a LED. If this LED is not glowing orange or red when the power is  
applied there is a severe problem. Either the module is broken or the necessary voltages for its use are not  
available.  
NOTE: For proper system operation when using the IC200, IC414, or other PCI extension interface to connect  
a CC10X crate to a remote PC, the crate must be powered on before the PC in order for the PC BIOS to  
recognize the presence of the CompactPCI crate.  
4. Devices that were installed using a previous version of Acqiris Software the instruments in these logical  
positions will still appear as Unknown Devices. This can be changed to the new Acqiris type category with  
the Grey Diamond icon by Uninstalling the device and then Installing again. Instructions on this procedure  
can be found in the ReadMe.txt file in the manuals folder of your Acqiris Software installation.  
2.5.  
After Restarting  
2.5.1. Windows 2000  
Under Windows 2000, you must login with administrator privileges after the first boot following the hardware  
installation; the Plug&Play system must have the appropriate privileges to be able to complete your hardware  
installation successfully. After a successful hardware installation, you will be able to use your Acqiris Digitizer(s)  
with normal privileges.  
At the first boot following the hardware installation, Windows will detect the new hardware and will install the  
devices automatically. The following image will appear.  
NOTE: In some systems an application program (such as AcqirisLive) will not yet work correctly at this point. One  
additional boot cycle may be needed if this is the first time that a hardware board is being installed.  
2.5.2. Windows XP  
Under Windows XP, you must login with administrator privileges after the first boot following the hardware  
installation; the Plug&Play system must have the appropriate privileges to be able to complete your hardware  
installation successfully. After a successful hardware installation, you will be able to use your Acqiris Digitizer(s)  
with normal privileges.  
If you login with administrator privileges after the first boot following the hardware installation, Windows will detect  
the new hardware and start the “Found New Hardware Wizard” after a few seconds.  
User Manual: Family of 10-bit Digitizers  
Page 23 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
To proceed with the installation click "No, not this time" and then “Next”. The following window will appear:  
To proceed with the installation click “Next”. You should then see  
User Manual: Family of 10-bit Digitizers  
Page 24 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
and then  
NOTE: In some systems an application program (such as AcqirisLive) will not yet work correctly at this point. One  
additional boot cycle may be needed if this is the first time that a hardware board is being installed.  
User Manual: Family of 10-bit Digitizers  
Page 25 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2.6.  
LabVIEW RT  
During program development you can choose whether you use LabVIEW or LabVIEW RT compatible libraries by  
switching the version present in National Instruments\LabVIEW m.n\instr.lib\. This swap can be facilitated by  
using the Install VI library for LabVIEW or LabVIEW Real-Time shortcut available in the Shortcut folder under  
Start Programs.  
There is only one Acqiris Driver. It supports all Acqiris Instruments. The instructions below concern LabVIEW RT  
as used in NI PXI processors.  
The Aq_RT.inf and AqRT_4.ini files must be uploaded to the target. To do this,  
start the MAX application,  
right click on the target  
select file transfer  
select the Aq_RT.inf file on your host machine and upload ('To Remote') to the LabVIEW RT working  
directory (/NI-RT/system) on the target  
select the Aq_RT.ini file on your host machine and upload ('To Remote') to the LabVIEW RT working  
directory (/NI-RT/system) giving it the name AqDrv4.ini  
For Acqiris modules which need FPGA files you should,  
create the folder \firmware in the /NI-RT/system directory using the file transfer application  
select the FPGA files (from <AcqirisDxRoot>\Firmware) you want to copy to the target and upload  
them into the firmware directory  
Restart the target after finishing the file transfers.  
Restart the MAX Explorer and you should have Acqiris digitizers detected in your PXI system.  
2.7.  
Installing the IVI-COM/C Driver  
Please install Acqiris Software for Windows first. Then the Acqiris Software CD Autorun program gives access to  
two installers:  
ƒ
ƒ
IVI Shared Components 1.4  
IVI-COM/C IviAqD1 driver  
These must be installed in the order shown above. For more information you can then consult the Readme.txt file in  
the IVI\Drivers\IviAqD1 folder or the documentation through the program shortcut present under Ivi/IviAqD1.  
2.8.  
Distribution for Windows 2000/XP and Linux  
The manuals/ReadMe.txt file contains a list of files to be found after a complete installation of Acqiris Software on  
Windows systems. Similarly the ReadMeLinux file gives the list of files corresponding to that installation.  
User Manual: Family of 10-bit Digitizers  
Page 26 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
3.  
Product Description  
3.1.  
Overview  
Simplified Block Diagram  
once for each channel  
Signal  
Input  
10-bit  
10- bit  
DEMUX  
256K  
Acq Mem  
Input  
Signal  
Amplifier  
SH + ADC  
50  
Ohms  
TIMEBASE  
trigger  
Input  
Trigger  
Signal  
Amplifier  
TRIGGER  
circuit  
Thr  
DAC  
50  
Ohms  
PCI Interface  
Cal  
DAC  
Card  
Controller  
PCI Bus  
The 10-bit DC series products are CompactPCI/PXI compliant and require an appropriate CompactPCI crate. The  
digitizers are all fully programmable over the PCI bus and deliver oscilloscope-like performance. Data captured by  
the digitizers can be transferred to a host processor, either in the PC, in the crate, or interfaced to the crate, over the  
PCI bus. The 6U units (DC222, DC252, and DC282) can operate at 64-bit 66 MHz with peak transfer speeds  
approaching 500 MB/s. The 3U units (DC122 and DC152) work at 32-bit 33 MHz. The digitizers occupy a single  
slot of a 3U or 6U CompactPCI crate and they comply with the 3U or 6U CompactPCI standard (PXI compliant).  
Acqiris digitizers are designed to provide superior measurement precision and accuracy. Key acquisition  
specifications (such as DC accuracy, integral and differential non-linearity) have been optimized to deliver maximum  
measurement fidelity. Careful circuit layout, custom IC’s and special packaging techniques have all been employed  
to reduce overall system noise. The use of custom IC’s also dramatically reduces the total number of discrete  
components required. This has tremendous benefits on reliability and also allows the modules to use a minimal  
amount of power.  
For complete technical specifications concerning your particular digitizer please refer to the product’s Specifications  
& Characteristics. In addition, we maintain up-to-date versions of all product data sheets on our web site  
you have trouble accessing our web site, or viewing the data sheets, please contact your nearest sales office.  
User Manual: Family of 10-bit Digitizers  
Page 27 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
3.2.  
Channel Input  
The 10-bit digitizer family can be configured with a variety of front-end mezzanine boards offering different  
capabilities. The currently available front end options are detailed below:  
Impedance Bandwidth  
Guaranteed  
BW Limiter  
selections  
FS Voltage  
Ranges  
Maximum  
Offset  
/ Max. Voltage  
Front End  
Option  
(typical)  
/ Risetime  
Standard  
(Std)  
2 GHz  
0.17 ns  
20, 200, 700 MHz  
20, 200, 700 MHz  
0.05 – 5.0 V  
0.05 – 5.0 V  
50 Ω  
50 Ω  
± 5 V  
± 5 V  
± 5 V  
± 5 V  
950 MHz  
( 1 GHz)  
0.35 ns  
High-  
Impedance  
(HZ)  
(300 MHz)  
1.2 ns  
20, 200, 700 MHz  
(see remark below)  
0.05 – 50 V  
1 V  
1 MΩ  
50 Ω  
± 20 V (> 0.5 V FS)  
± 200 V (> 5V FS)  
± 300 V  
3 GHz  
0.12 ns  
± 0.5 V  
High-  
Frequency  
± 2 V  
(HF)  
A given unit’s model number will indicate which front end option is installed. If nothing follows the number itself  
then the standard front end option is present.  
Units with either Standard or HF front end options can be ordered with BNC or SMA connectors. The HZ units use  
BNC connectors.  
For FS > 5 V only the 20 MHz Bandwidth Limiter is available.  
3.2.1. Standard & High Impedance Coupling  
Both AC and DC coupling modes are available. The AC mode couples signals capacitively thus removing the input  
signal’s DC component and filtering out any signal component below 16 Hz for the HZ option or 32 Hz for the Std  
option. DC mode allows all signal components to be passed through to the digitizer.  
3.2.2. Impedance  
The input channels of the 10-bit Digitizer Family offer termination in either 50 Ω or 1 MΩ. The 50 Ω coupling mode  
for the Standard and High-Impedance front ends offers high quality termination with better than ± 1% precision. It is  
ideally suited to use with 50 Ω transmission lines (coax), high bandwidth low impedance (typically 500 Ω) probes or  
active probes. The HF 50 Ω coupling mode has ± 2% precision The 1 MΩ coupling mode provides a high impedance  
(low load) capability that is suited for use with most standard high impedance probes. The high impedance mode also  
features low (14 pF typical) capacitance that helps to minimize loading effects that can occur when probing high  
frequency circuits.  
3.2.3. Input Protection  
The input amplifiers are protected against over-voltage signals as shown in the table. The High-impedance limit  
refers to the input signal’s DC + peak AC < 10 KHz voltage.  
3.2.4. Mezzanine Front-end  
The front-end electronics are all mounted on a removable mezzanine card. In the event of accidental damage or as  
components fatigue over time (e.g. relays in high duty cycle automated testing applications), the mezzanine card  
allows for fast and efficient replacement.  
User Manual: Family of 10-bit Digitizers  
Page 28 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
3.2.5. Bandwidth and Rise Time  
The bandwidth specification indicates the frequency at which an input signal will be attenuated by 3 dB  
(approximately 30% loss of amplitude). The bandwidth also affects the minimum rise and fall times that can be  
passed through the front-end electronics. A pulse with a very sharp edge will be observed to have a minimum rise  
time (τmin) determined by the front-end electronics. In general a pulse with a given 10-90% rise time (τ10-90real) will be  
observed with a slower value given by:  
2
τ10-902 = τ10-90real2 + τmin  
where τmin (ns) 0.35 (GHz-ns) / BW (GHz)  
For the HZ option in 1 MΩ coupling the 700 MHz Bandwidth Limiter cannot be used. Furthermore, for FS gain >  
5V the 200 MHz Bandwidth Limiter is always active.  
3.2.6. Input Voltage and Offset  
The input channel provides a fully programmable amplifier with variable input voltage and offset. Full Scale (FS)  
input voltages are selectable in a 1, 2, 5 sequence for the range shown above. Care should be taken to select an input  
voltage range that will allow the signal to be recorded using as much dynamic range of the digitizer as possible. The  
Variable Offset is programmable in the range of ±2 V when using an FS Input Voltage setting of 500 mV or below,  
increasing to ± Maximum Offset for FS settings above 500 mV (see the preceding table). The raw 10-bit ADC data  
values are read as 16-bit integers in the range [-32768, +32704] with nominal steps of 64 LSB’s between successive  
codes and the first and last values reserved for underflow and overflow respectively. The midpoint value, 0, of the  
range corresponds to the negative of the offset voltage. Thus the Full Scale Range (FSR) goes from  
–Offset Voltage  
(FS/2)  
to  
–Offset Voltage  
+
(FS/2)  
Signals going outside of the FSR will be clipped and data values for the clipped portion of a signal should be  
regarded as erroneous.  
3.2.7. Vertical Resolution  
The digitizers described in this manual use an ADC system with 10 bits of vertical resolution (1024 levels). The  
dynamic range of the ADC covers the Full Scale Range (FSR) of the Input Voltage setting. For example, if the Input  
Voltage is set to 1 V then the ADC resolution is equivalent to 0.977 mV. To obtain the best dynamic range from the  
ADC care should be taken to ensure that the input signal varies over more than 50% of the Input Voltage FSR  
setting.  
3.2.8. DC Accuracy and Linearity  
These 10-bit digitizers use low noise front-end electronics in order to ensure voltage measurement is made with  
accuracy and precision. DC voltage accuracy, at 0 V offset, is better than ±2% (±1% typical) of the input voltage full  
scale. The static differential nonlinearity is < 2 LSB. The integral linearity is good to ± 1 LSB (typical).  
3.2.9. Using Probes  
The 50 Ω and 1 MΩ input impedance settings make it possible to use Acqiris digitizers with a wide variety of  
probes. The 50 Ω setting is most commonly used for active probes and low impedance (500 Ω) passive probes. The  
1 MΩ setting is normally used for high impedance probes. While it is possible to use a variety of different  
manufacturers high impedance probes we recommend the use of our model P001 Passive Probe. The model P001 is  
matched to the Acqiris digitizers to ensure that effects that can reduce measurement quality, such as bandwidth loss  
and capacitive loading, are minimized. The P001 features 300 MHz bandwidth, 10 MΩ input impedance and 10:1  
attenuation. These specifications make it an excellent choice for most general purpose probing applications. Before  
using any Passive Probe with a digitizer care should be taken to check that the probe has been correctly adjusted  
(refer to the Probe’s Calibration procedure).  
NOTE: Passive high impedance probes are not suitable for high fidelity measurements above 100 MHz. The non-  
negligible (5-10 pF) tip capacitance loads the signal causing distortion and/or ringing when combined with the  
ground lead inductance.  
User Manual: Family of 10-bit Digitizers  
Page 29 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
3.3.  
Data Acquisition  
The table below summarizes the characteristics discussed in the sections that follow:  
Model  
Agilent #  
Max.  
Sampling  
Rate  
Max.  
Default Memory  
points/ channel  
Maximum  
Optional  
Memory  
Maximum  
Segments  
CONVERTERS  
PER CHANNEL/  
CHANNELS  
points/ channel  
DC122  
U1062A  
DC152  
U1062A  
DC222  
U1065A  
DC252  
U1065A  
DC282  
U1065A  
4 GS/s  
4 GS/s  
8 GS/s  
8 GS/s  
8 GS/s  
1 / 1  
2 / 2  
1 / 1  
2 / 2  
4 / 4  
512K  
256K  
1024K  
512K  
256K  
--  
1K  
--  
1K  
1G  
125K  
125K  
125K  
512M  
256M  
3.3.1. Sampling Rate  
All Acqiris digitizers contain an analog-to-digital conversion (ADC) system that can sample waveforms, in a real  
time sampling mode, at rates from the maximum allowed rate down to 100 S/s (10 ms per point). The sampling rate  
can be programmed and is selectable in a 1, 2, 2.5, 5 sequence (i.e. 1 MS/s, 2 MS/s, 2.5 MS/s, 5 MS/s, 10 MS/s, … 1  
GS/s, 2GS/s, 4 GS/s, 8 GS/s). The maximum sampling rate shown above exploits the possibility of combining  
channels.  
3.3.2. Acquisition Memory  
Data from the ADC is stored in on-board acquisition memory. The amount of memory in use for acquisition can be  
programmed and is selectable from 1 point to the full amount of acquisition memory available.  
For technical reasons, a certain acquisition memory “overhead” is required for each waveform, reducing the available  
memory by a small amount. In order to simplify programming, an interface function recommends the best sampling  
rate and the maximum possible number of data points, taking into account the available memory, the requested time  
window, the number of segments (in Sequence mode), as well as the required memory overhead.  
To ensure maximum sampling rate and high timing resolution, we strongly recommend the use of long acquisition  
memories whenever possible.  
Optional Memory is available on most models. It allows the maximum number of segments to be increased  
significantly. However, this memory is divided into 2K sample pages and a segment must start at the beginning of  
the page. Additional memory is used for segment time stamps. These constraints must be kept in mind when trying to  
understand how the driver limits the number of points as the number of segments changes. Furthermore, multi-  
segment event readout must read entire pages for each ADC. This can introduce significant overhead for small  
segments and/or when channels are combined.  
3.3.3. Single and Sequence Acquisition Modes  
Digitizers acquire waveforms in association with triggers. Each waveform is made of a series of measured voltage  
values (sample points) that are made by the ADC at a uniform clock rate. The digitizer can also measure and store  
the arrival time of each trigger using the information from the on board Trigger Time Interpolator (TTI). Readout of  
the individual trigger time stamps makes it possible to determine the time from one trigger to any other trigger. Time  
differences up to 213 days can be measured. The TTI resolution sets the resolution of the trigger time stamps (see  
section 3.3.5 TIMING). To maximize sampling rates and utilize memory as efficiently as possible, the digitizers  
include both Single and Sequential storage modes. For both of these modes the data of all of the active channels is  
acquired synchronously; all of the ADC’s are acquiring data at the same time, to within a small fraction of the  
maximum sampling rate.  
User Manual: Family of 10-bit Digitizers  
Page 30 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
The Single Acquisition mode is the normal operation of most digitizer products. In this mode an acquisition consists  
of a waveform recorded with a single trigger. The user selects the sampling rate and acquisition memory size and sets  
the number of segments to 1 (default value).  
The Sequence Acquisition mode allows the capture and storage of consecutive “single” waveforms. Sequence  
Acquisition mode is useful as it can optimize the digitizer’s sampling rate and memory requirements for applications  
where only portions of the signal being analyzed are important. The mode is extremely useful in almost all impulse-  
response type applications (RADAR, SONAR, LIDAR, Time-of-Flight, Ultrasonics, Medical and Biomedical  
Research, etc.).  
In Sequence Acquisition mode the acquisition memory is divided into a pre-selected number of segments.  
Waveforms are stored in successive memory segments as they arrive. Each waveform requires its own individual  
trigger. The memory can be divided into any number of segments between 2 and 1000 (up to 16000 segments with  
the M32M option in a DC282 and 125000 segments with options M256M or greater). In Sequence Acquisition mode  
the user needs to specify the sampling rate, the total acquisition memory, and the number of segments. Note that the  
Single Acquisition mode is just a special case of the Sequence Acquisition mode with the number of segments set to  
1.  
Sequence acquisition enables successive events, which can occur within a very short time, to be captured and stored  
without loss. A crucial feature of Sequence Acquisition mode is that it has a very fast trigger rearm time. A fast  
trigger rearm helps produce very low “dead time” (< 350 ns for the highest available sampling rates using internal  
memory and < 1.8 μs for the highest available sampling rates with optional memory) between the segments of a  
sequence acquisition. The “dead time” is the period after the end of an event when the card cannot digitize data for a  
new trigger event.  
3.3.4. Simultaneous Multibuffer Acquisition and Readout (SMAR)  
The internal memory has a dual-port structure which can be exploited to permit simultaneous data acquisition and  
read out. The memory can be turned into a circular buffer of a chosen number of banks, between 2 and 1000. Data  
can be read out of one bank while data is acquired into any available free banks. This mechanism, together with  
sequence acquisition, helps achieve a high maximum continuous event rate in spite of interrupts due to the computer  
operating system. The maximum continuous event rate is the maximum value of the trigger frequency that can be  
accepted without the loss of any event. Note that this mode is available for internal memory only and does not work  
for all cases of sampling rate, channel combination, and ASBus2 usage. It will work for the maximum sampling rate.  
This mode is available for the DC222, DC252, and DC282 only. Please contact us if you have any doubts.  
3.3.5. Timing  
A crystal-controlled timebase is used to clock the ADC system of the digitizers. The timebase accuracy is guaranteed  
to be better than 2 ppm . The digitizers also include a built-in Trigger Time Interpolator (TTI) that measures the time  
from the trigger point to the first sample point. This information is essential for determining the precise relation  
between the trigger or other event of interest and the digitized samples of the signal. The TTI resolution is ~10 ps.  
3.3.6. Timebase Range  
The timebase range defines the time period over which data is being acquired. For example, the DC282 has a  
standard acquisition memory of 256 Kpoints and maximum sampling rate of 2 GS/s. Therefore, at the maximum  
sampling rate, the digitizer can record a signal over a timebase range of up to 130 μs (approx. 260,000 points *  
0.5 ns/point). The timebase range can be adjusted by varying the amount of acquisition memory or the sampling rate  
of the digitizer.  
3.3.7. Combining channels  
The DC152, DC252, and DC282 digitizers offer the possibility of combining the converters (and their memories)  
from two or four channels to analyze a single input channel. With this feature the maximum sampling rate and the  
maximum amount of acquisition memory can be doubled or quadrupled if all of the input channels are not of  
immediate interest.  
3.3.8. Data Readout  
The DC222, DC252, and DC282 digitizers are capable of handling 64-bit 66 MHz readout. Such operation requires a  
single board computer or a cPCI interface + PC supporting such functionality. All devices in the crate must be  
capable of 66 MHz readout. The crate may also have to be set so as to allow this mode; this means M66EN must be  
open and V(I/O) should be 3.3 V. Acqiris CC103, CC105, and CC108 crates have toggle switches to select such use.  
User Manual: Family of 10-bit Digitizers  
Page 31 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
This cannot be done in larger crates such as the Acqiris CC121. The actual transfer speed obtainable will depend on  
many other system characteristics.  
3.4.  
Trigger  
Normally the trigger settings applied to the digitizer are used to determine the time at which the device will stop  
acquiring data. They are also capable of a ‘Start on Trigger’ mode of acquisition (see the Programmer’s Guide for  
further details). The various trigger settings are outlined below.  
3.4.1. Trigger Source  
The trigger source can be a signal applied to either an Input Channel (internal triggering) or the External Trigger  
Input. A standardized trigger in signal can also be routed via the PXI Bus Star Trigger line. Because of their high  
bandwidth HF front ends do not allow internal triggering.  
The digitizers provide a separate front panel input that can be used as an External Trigger Input. The External Input  
provides a fully functional trigger circuit with selectable coupling, level and slope as for the Internal Triggering  
source. The external trigger has a fixed 50 Ω termination impedance. It allows optional BW limiter selections of 20,  
200, or 700 MHz. The external trigger circuit has diode protection against overload. In all 50 Ω cases a ±5 V limit on  
trigger signals should be respected, although somewhat higher voltages for short time periods will not damage the  
unit.  
3.4.2. Trigger Coupling  
Trigger coupling is used to select the coupling mode applied to the input of the trigger circuitry. Modes available  
include AC LF Reject and DC. The AC LF Reject mode couples signals capacitively and removes the input signal's  
DC component and signals below 50 Hz. DC mode allows all signal components to be passed through to the trigger  
circuit. The HF Reject mode removes signal components above 50 KHz.  
3.4.3. Trigger Level  
The trigger level specifies the voltage at which the selected trigger source will produce a valid trigger. The trigger  
level is defined as a set voltage. Using the internal trigger, the level is set with respect to the midpoint voltage (Vm= –  
Offset voltage) of the digitizer’s vertical scale. Internal trigger level settings (expressed in %) must be within Vm ±  
0.5 FS, where FS is the channel Full Scale. All trigger circuits have sensitivity levels that must be exceeded in order  
for reliable triggering to occur.  
The digitizers allow the user to choose the external trigger Full Scale from the set of values 0.5, 1.0, 2.0 or 5.0 V.  
The external trigger level can then be set to values in the range ± 0.5 FS.  
The digitizers will trigger on signals with a peak-peak amplitude > 15% FS from DC to their bandwidth limit.  
3.4.4. Edge Trigger Slope  
The trigger slope defines which one of the two possible transitions will be used to initiate the trigger when it passes  
through the specified trigger level. Positive slope indicates that the signal is transitioning from a lower voltage to a  
higher voltage. Negative slope indicates the signal is transitioning from a higher voltage to a lower voltage.  
3.4.5. Window Trigger  
The digitizers implement a Window trigger. Two trigger level thresholds are used to define the desired range. The  
trigger can then be chosen to occur either when the signal exits or enters the window range. This mode can be  
thought of as the appropriate OR of two edge triggers of opposite slope.  
3.4.6. HF Trigger  
The digitizers implement an HF trigger that allows triggers to be reliably accepted at rates above 1 GHz. In this  
mode, triggers occur on every fourth positive edge. The window trigger mode is not available.  
3.4.7. Spike Stretcher  
The trigger circuit also has a Spike Stretcher mode which ensures that even very short pulses are capable of  
generating triggers. This mode is useful if the time interval during which the trigger signal satisfies the threshold  
condition is less than 0.5 ns and the trigger frequency is less than 10 MHz. The trigger slope is positive in this mode.  
User Manual: Family of 10-bit Digitizers  
Page 32 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
3.4.8. Pre- and Post-Trigger Delay  
To increase trigger flexibility a pre- or post-trigger delay can be applied to the trigger position.  
The amount of pre-trigger delay can be adjusted between 0 and 100% of the acquisition time window (i.e. sampling  
interval x number of samples), whereas the post-trigger delay can be adjusted within the time interval corresponding  
to [0, 235 – 1 samples].  
Pre- or post-trigger delays are just different aspects of the same trigger positioning parameter:  
The condition of 100% pre-trigger indicates that all data points are acquired prior to the trigger, i.e. the trigger  
point is at the end of the acquired waveform.  
The condition of 0% pre-trigger (which is identical to a post-trigger of 0) indicates that all data points are  
acquired immediately after the trigger, i.e. the trigger point is at the beginning of the acquired waveform.  
The condition of a non-zero post-trigger delay indicates that the data points are acquired after the trigger occurs,  
at a time that corresponds to the post-trigger delay, i.e. the trigger point is before the acquired waveform.  
The digitizer hardware accepts pre- and post-trigger adjustments in increments of 16 samples. By definition post-  
trigger settings are a positive number and pre-trigger settings are a negative number.  
Thus it is only natural that the software drivers treat pre- and post-trigger delays as a single parameter in seconds that  
can vary between –nbrSamples * samplingInterval (100% pre-trigger) and +maxPostTrigSamples * samplingInterval  
(max post-trigger). Since the Acqiris software drivers provide very accurate trigger position information upon  
waveform readout, the accepted resolution of the user-requested pre-/post-trigger delay is much better than 16  
samples. For more details, refer to the Programmer’s Guide.  
3.4.9. Trigger Status  
The front panel includes a tri-color LED indicator to show the status of the trigger. When the LED is green it  
indicates the trigger is armed and waiting for a valid trigger to occur. Red indicates that the trigger has occurred, the  
acquisition is complete and the data is waiting to be readout. The user can override the default functions and program  
the LED color in an application-specific manner.  
3.5.  
External Clock and Reference  
For applications where the user wants to replace the internal clock of the digitizer and drive the ADC with an  
external source, either an External Clock or an External Reference signal can be used. The Clock or Reference  
signals can be entered into the digitizer by the dedicated MMCX. In addition, the PXI Bus 10 MHz system clock  
signal (PXI_CLK10) can be used as the reference.  
The External Clock must be continuously present for use by these digitizers. The input signal must have a frequency  
between 200 MHz and 2 GHz and a minimum amplitude of at least 0.5 V peak to peak into 50 Ω at the front of the  
digitizer. The transitions of the clock are defined with the aid of a threshold that is user selectable in the range [-3.0  
V, 3.0 V]. The signals should not exceed ±5 V amplitude. For a detailed discussion on the programmed use of the  
external clock, refer to the Programmer’s Guide.  
For applications that require greater timing precision and long-term stability than is obtainable from the internal  
clock, a 10 MHz Reference signal can be used. The External Reference is nominally at 10 MHz. However,  
frequencies in the range [9.97 MHz, 10.03 MHz] will be accepted. If you do this you may need to correct for the  
difference in your application since the digitizer and the driver have no way to know about such deviations. The  
amplitude and threshold conditions, for an External Reference, are the same as for the External Clock. If  
synchronization between several digitizers is required, the reference signal should be applied to all of them.  
3.6.  
Internal Calibration  
The software drivers supplied include calibration functions for the timing, gain and offset settings, which can be  
executed upon user request. The digitizers are never calibrated in an “automatic” way, i.e. as a side effect of another  
operation. This ensures programmers have full control of all calibrations performed through software in order to  
maintain proper event synchronization within automated test applications.  
The digitizers include a high precision voltage source and a 16-bit DAC, used to determine the input voltage and  
offset calibration.  
For accurate time and voltage measurements it is recommended to perform a calibration once the module has attained  
a stable operating temperature (usually reached within 15 minutes of digitizer operation after power on). Further  
calibration should not be necessary unless temperature variations occur.  
User Manual: Family of 10-bit Digitizers  
Page 33 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
A full internal calibration of a digitizer can be very time consuming (> 100 s/digitizer), in particular for the HZ  
models. Therefore, several other options are available. They are documented in the Programmer's Reference  
Manual. A program can always be started with the digitizer in an uncalibrated state and data taken can be used for  
many kinds of testing. However, as soon as good data respecting the specifications of the instrument is required a  
calibration of at least the current acquisition state is needed. The full internal calibration has the advantage that it  
generates the calibration constants needed for any possible configuration of the instrument; its disadvantage is the  
time taken. If a more selective calibration is done it will allow the generation of good data in the current acquisition  
state. This calibration will remain useable whenever that acquisition state is used again for as long as the temperature  
of the instrument does not change significantly. A fast calibration of a channel in a configuration can be done in  
around a second. Many applications can save time by only performing calibration for the configurations that will  
actually be used.  
An application program will close devices if it exits normally. When a 10-bit digitizer is closed in this way its power  
consumption will be reduced. This has the consequence that if another application program is started later an  
appropriate warm-up period may have to be respected before a calibration is called for and full performance  
expected.  
3.7.  
ASBus2  
The digitizers may be used in applications that require many data acquisition channels. In such cases it is possible to  
use more than one digitizer in a standard CompactPCI/PXI crate. Each DC Series 10-bit digitizer includes ASBus2, a  
proprietary high bandwidth auto-synchronous bus system that allows multiple digitizers to work together  
synchronously.  
ASBus2 distributes both the clock and trigger signals along a plug-in front panel bus, between all the digitizers that  
participate in the system. It allows any digitizer to act as the trigger source and any other digitizer to act as the clock  
source (acquisition master), enabling all the digitizers to be clocked at the same time. Synchronizing the clock signals  
between the devices improves the accuracy of cross-channel measurements and is essential for accurate time  
correlation between signals on different channels.  
ASBus2 is intended to connect modules of the same type, i.e. of the same model number, although some exceptions  
to this rule might be possible. If modules with the same model number, but different memory length options, are  
connected only the shortest memory length can be used.  
The ASBus2 connector is located on the front panel of each module. Bridges are used to connect adjacent modules  
for synchronization. An ASBus2 multi-instrument is activated through software by a function call. Up to 5 6U  
modules, or 3 3U modules, may be synchronized with the ASBus2. More detailed information and the commands  
required to set up the ASBus2 clock and trigger distribution are included in the Programmer's Guide and Reference  
manuals.  
3.8.  
Special Front Panel Inputs and Controls  
Connector  
MMCX CLK IN  
MMCX I/O A  
Function  
50 External Reference or External Clock Input  
User configurable (see below)  
MMCX TR OUT Signal occurs after an accepted TRIGGER.  
MMCX I/O B User configurable (see below)  
The I/O A and I/O B signals are 3.3 V compatible CMOS. This means that, on input, low is < 0.7 V and high must be  
in the range [1.7 V, 5.0 V]. An unconnected signal will be high. This definition ensures TTL compatibility. On  
output, the low level will be in the range [0 V, 0.7 V] and the high level in the range [1.7 V, 3.3 V] for HiZ. The high  
level output will typically give 0.8 V into 50 Ω. The I/O A or B connectors can be used for an Enable trigger input or  
the following output signals: 10 MHz reference clock, Acquisition skipping to next segment, Acquisition is active,  
Trigger is armed.  
3.9.  
External Trigger Output  
When the digitizer is ready to be triggered and a valid trigger signal occurs, a trigger output is generated for external  
use. It is always available on the Front Panel Trigger Out MMCX connector. The pulse ends when the data  
acquisition for the trigger in question is complete.  
User Manual: Family of 10-bit Digitizers  
Page 34 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
NOTE: The External Trigger Output functionality is implemented in the hardware. No Trigger Out signal occurs for  
software-generated triggers such as those of the AUTO mode of AcqirisLive or through the use of the function  
AcqrsD1_forceTrigger. However, AcqrsD1_forceTriggerEx does generate the signal.  
Signal: 1.6V swing  
Trigger Output Block diagram:  
50 Ohm  
0v centered  
Trigger Out  
G=1  
The output swing is 1.6 V (±  
when unloaded and 0.8 V when  
terminated on 50 Ω. The rise  
0.8  
V)  
Offset: +/-2.5 V  
and fall  
times are 2.5 ns typical. The offset can be adjusted, by software control in the range [–2.3 V, +2.3 V] unloaded, or  
[-1.15 V, +1.15 V] into 50 Ω. The maximum output current capability is ± 15 mA. As the output is retro-terminated,  
it is possible to drive a 50 Ω line unterminated (HiZ)  
+5V  
without loss of performance.  
ECL  
Signal  
TTL  
signal  
R
220  
R
68  
For a TTL compatible signal, set the offset to 1.0 V and the  
at destination will be +0.2 to +1.8 V.  
swing  
Trigger  
Out  
With  
Offset  
-1.18V  
Trigger Out  
With  
Offset  
R
For an ECL compatible signal, terminated on 50 Ω to –1.2  
the offset to –1.2 V and the output will be in the range [–0.8  
V]).  
V, set  
V, –1.6  
R
68  
220  
+1.18V  
-5V  
Alternatively, to reduce the current drawn from the digitizer,  
terminations shown here can be used:  
the  
A standardized trigger out signal can also be routed to the PXI Bus Star Trigger line.  
3.10. Electrical, Environmental and Physical Specifications  
3.10.1. Electrical  
PCI  
Max. Power  
Current Requirements (A)  
Rev.  
Consumption (W)  
Model  
+12 V  
+5 V  
+3.3 V  
-12 V  
DC122  
Default  
Memory  
2.2  
31  
40  
34  
43  
62  
80  
0.1  
3.3  
3.4  
0.03  
DC122  
Optional  
Memory  
2.2  
2.2  
2.2  
2.2  
2.2  
0.1  
0.1  
0.1  
0.1  
0.1  
3.3  
3.9  
3.9  
6.7  
6.7  
6.0  
3.4  
0.03  
0.03  
0.03  
0.03  
0.03  
DC152  
Default  
Memory  
DC152  
Optional  
Memory  
6.0  
DC2x2  
Default  
Memory  
7.2  
DC2x2  
12.5  
Optional  
Memory  
The Maximum Power Consumption has been increased by 5% over the value calculated with the currents shown to  
take into account higher allowed values of the crate voltages. The current requirements of the DC222 and DC252  
models are only slightly lower than the DC282 case. The differences between the different front end options are also  
small; a single Standard or HZ channel pulls ~0.3 A @ +5 V.  
The electrical requirements shown above can impose loading limitations in the cPCI crate. In the extreme case where  
a PC503 processor is present in an Acqiris crate the following limits apply:  
User Manual: Family of 10-bit Digitizers  
Page 35 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Crate  
DC1x2  
DC2x2  
DC2x2  
all models Default  
Memory  
Optional  
Memory  
CC103 /  
CC105  
None  
None  
None  
CC108  
None  
4
4
normal and  
low power  
CC108-600 None  
CC121  
None  
7
4
9
15  
If a third party crate is being used please check that the current drawn at 3.3 V does not sag below the effective PCI  
allowed tolerance at the beginning of the acquisition.  
The digitizers can use the PCI Bus at 32-bit 33 MHz, all models, or at 64-bit 66 MHz, DC2x2 only . The latter  
allows data to be transferred with DMA at average rates up to 350 MB/s with appropriate hardware under Windows  
and Linux. Otherwise, the DMA transfer rates are limited to ~100 MB/s. The CompactPCI modules are compatible  
for either V I/O = 3.3 V or 5V.  
3.10.2. Environmental and Physical  
The modules have a Declaration of Conformity according to ISO/IEC Guide 22 and EN45014 and CE Marks of  
Compliance.  
The American Standard MIL-PRF-28800F has been used as the source for many of the characteristics specified  
below. Those indicated with a ‘*’ are compatible with Class 3 as defined in that standard.  
The front panels of the CompactPCI modules are in compliance with the IEEE 1101.10 standard.  
Operating Temperature  
0° to 40°C  
The above values are for the ambient temperature of the room (or equivalent) where the CompactPCI crate is located.  
The temperature as measured on the board may well be significantly higher. On-board temperatures above 60°C  
should be avoided.  
Relative Humidity*  
5 to 95% (non-condensing)  
Dimensions  
The DC122 and DC152 conform to the 3U CompactPCI standard (100 mm × 160 mm × 20 mm). All other DC  
modules conform to the 6U CompactPCI standard (233 mm × 160 mm × 20 mm).  
Shock*  
30 G, half-sine pulse  
Vibration*  
5-500 Hz, random  
Safety  
Complies with EN61010-1  
EMC Immunity  
User Manual: Family of 10-bit Digitizers  
Page 36 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Complies with EN61326-1: Industrial Environment  
EMC Emissions  
Complies with EN61326-1: Class A for radiated emissions  
Required Airflow  
> 2 m/s in situ  
User Manual: Family of 10-bit Digitizers  
Page 37 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4.  
Running the AcqirisLive Application  
AcqirisLive is an application to control and demonstrate the capabilities of Acqiris Digitizers on a single machine.  
AcqirisMAQS is a more advanced application offering many interesting possibilities for the control of acquisition  
systems in a single or multi-machine environment. Ask your salesman or Acqiris for more information.  
4.1.  
Getting Started with AcqirisLive  
Once the software and hardware installation described in Section 2 of this manual is complete, you will find  
AcqirisLive in the start menu of your computer. Start AcqirisLive from the menu. An initial AcqirisLive load screen  
will appear followed by a panel of controls and a waveform display window.  
During startup, AcqirisLive searches for all Acqiris Digitizers on the PCI bus. If no digitizers are found on the PCI  
bus, AcqirisLive will display an error message indicating that no digitizers were found, and the program will  
continue in simulation mode. If this is not wanted, turn off the computer, install and turn ON the hardware on the PCI  
bus and, then, restart the computer.  
AcqirisLive is intended to verify the functionality of the card and to serve as an easy to use and convenient program  
for capturing, displaying, and archiving captured data into the format of your choice, binary or ASCII. AcqirisLive is  
provided free with the purchase of any Acqiris Digitizer.  
We supply drivers on the CDROM for National Instruments LabWindows/CVI, LabVIEW, Microsoft Visual C++,  
and Visual Basic. We also provide some examples (GetStarted applications) as a guide or starting point for  
programmers developing their own applications to interface with the DC series digitizer cards.  
When you start AcqirisLive a display like that shown below appears.  
AcqirisLive displays two independent windows, a control window and a display window. The control window  
contains functions that allow you to manipulate the acquisition parameters of the card. In order to display an acquired  
waveform, connect a signal to the input, and then select the Auto acquisition mode at the bottom of the control panel.  
A waveform should appear in the display.  
Other controls in the control panel allow you to select the optimal acquisition settings for your particular signal. Each  
of the other control panel functions will be discussed in detail in the following Control Panel Functions section. Here  
is a sample result.  
User Manual: Family of 10-bit Digitizers  
Page 38 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
4.2.  
Control Panel and Functions  
4.2.1. Control Panel Mode  
DSO TR  
In AcqirisLive there are two control panel modes available, Oscilloscope Mode and Transient  
Recorder Mode. The choice of mode is entirely determined by the preference of the user.  
Both modes provide the same set of functional capabilities but display the settings slightly differently.  
For example in Transient Recorder Mode, the user selects the Full Scale Range of the amplifier whereas in  
Oscilloscope Mode the user selects the input voltage scale set in Volts/Div. The desired mode is selected using one  
of the two buttons shown at the top of the control panel. The distinction between the control panels for the two  
operating modes can be seen in the images below.  
In both modes it is possible to set the memory depth and the sample speed independently within the limits of the  
maximum available memory. However the two modes differ in the following way.  
In Oscilloscope Mode, the user selects the time window (time per div) and the sampling rate will be adjusted to the  
fastest possible within the limits of the available memory. The maximum available memory is programmable.  
In Transient Recorder Mode, the user selects the capture time window, the sampling rate, or the number of samples.  
When selecting the capture time window, the number of samples is adapted, keeping the current sampling rate and  
number of segments, up to the maximum available memory on the card. When selecting the sampling rate, the  
number of samples is adapted, keeping the current capture time window and number of segments, up to the  
maximum available memory on the card. When selecting the number of samples, the capture time window is  
adapted, keeping the current sampling rate and number of segments.  
User Manual: Family of 10-bit Digitizers  
Page 39 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Oscilloscope Mode  
Transient Recorder Mode  
4.2.2. Displaying Multiple Traces  
If multi-channel modules, or several single channel modules, are installed on the PCI bus, multiple waveform  
displays, one from each installed channel, are overlaid on the display.  
The current channel is selected by clicking on the channel name within the upper portion of the control panel. The  
vertical settings of the current channel and the timebase and trigger of the correspondent digitizer are then indicated.  
A unique name is given to each channel installed on the PCI bus. The current channel is indicated by a dark highlight  
over the name. Settings changed on the main control panel will only affect the active channel. Always verify proper  
acquisition setting for each channel prior to collecting data.  
There is a small rectangular indicator button to the left of each channel name in the channel list. If this indicator is  
filled in, the channel waveform is active; the channel is turned on and displayed in the display window. Clicking on  
this indicator once turns it transparent. This deactivates the reading of the data from the channel and clears the  
waveform display. Clicking once again on the indicator turns the channel back on and the indicator is filled in. After  
making such changes the channel data will not be displayed until after the next acquisition.  
The image at the left shows a channel list with four active channels  
in the system. All are turned on and displayed in the display  
window as indicated by the black indicators at the left. The DC271  
Ch1 is the currently selected channel.  
The channel list image at the right shows the same system with the  
DC271 Ch2 selected as the current channel. The other indicators are  
transparent and therefore their channels are turned off and not  
displayed.  
User Manual: Family of 10-bit Digitizers  
Page 40 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
As an alternative to controlling the channels to be displayed from the control  
panel you can also control them by selecting the desired channels from the  
list of channel gains shown at the right hand side of the display window.  
Pressing the Finder button at the top of the control panel will identify the  
module associated with the current channel. When the finder button is  
depressed, the LED on the front panel lights up yellow on the digitizer  
module with the current channel.  
4.2.3. Numeric Editor  
The numeric editor is shown in the control window at  
locations where numeric values are set. The Acqiris  
Numeric Editor enables easy input of numeric values,  
particularly those ranging over many decades. The numeric value can be  
changed in multiple ways.  
A number can be typed directly into the white number display area or  
alternatively, the Acqiris Numeric Editor may be used. In order to use the  
Acqiris Numeric Editor, click on the green numeric editor icon at the left of  
the number display. A numeric editor control will appear as shown.  
The methods to change the values are:  
1. Increment or decrement any digit with the arrow buttons by  
pressing the increment or decrement button. When the button is  
held down, the digit will continue to increment or decrement  
automatically.  
2. Click on a digit and drag the mouse up or down to increase or  
decrease the value.  
3. Type a new value with the keyboard for any highlighted digit.  
The value is constrained to fall within  
its allowed range whenever the user  
attempts to enter a larger or smaller  
value. Digits that are below the  
resolution or granularity cannot be  
modified. A message is displayed in  
the numeric setting control window  
when the pointer is positioned over a  
digit that is below the allowed granularity and cannot be changed.  
4.2.4. Keyboard shortcuts  
Two shortcuts have been defined to simplify applying changes of a parameter to a group of channels. When the  
mouse is over the control of a channel parameter -  
ƒ
ƒ
's' can be used to set the current value to all channels of all available instruments  
'c' can be used to set the current value to all channels of the current instrument.  
Other shortcuts are:  
ƒ
ƒ
ƒ
ƒ
'Ctrl-S' as a synonym for 'Save Setup'  
'Ctrl-R' as a synonym for 'Recall Setup'  
'Ctrl-C' as a synonym for Calibrate all (active) instruments'  
'Ctrl-E' functions as a toggle between 'Internal Clock' and 'External Clock' operation.  
4.2.5. Input Voltage Scale, Offset and Coupling  
Settings for input voltage scale, offset, coupling and impedance are  
available in the portion of the control panel shown to the right. The  
selected channel is shown on the upper section of the display. Only this  
channel will be affected by changes in any settings on the control panel.  
The input voltage scale is set by one of three possible methods:  
User Manual: Family of 10-bit Digitizers  
Page 41 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
1. Clicking on the white numeric display of the voltage scale will show a pop-up list. Select the desired gain  
setting with the mouse pointer.  
2. Clicking on the Decrement / Increment icons will step the voltage scale up or down to the  
next level for each click of the mouse.  
3. Clicking Max / Min will toggle to the least sensitive gain setting or to the most sensitive gain setting.  
The input Coupling can be set for AC or DC. The selected setting is shown as an actuated indicator with a green  
backlight. AC coupling blocks DC signal components from the input. The AC coupling mode acts as a high pass  
filter passing frequencies above 10 Hz. In DC coupling all frequency components are passed through the input.  
The input voltage offset control applies an offset to the input amplifier in order to better digitize and display signals  
offset away from 0 volts. Signals riding on a DC level can be most accurately characterized when the offset is  
adjusted so that the acquisitions are made about the midpoint of the digitizer’s Full Scale Range, and when as much  
of the 8-bit ADC range is used as possible. The ability to select an offset voltage enables accurate measurement when  
offsets are present in the signal.  
The Impedance selection allows the user to choose the appropriate internal input termination of 50 Ω or 1 MΩ.  
The Adv button in this section gives access to the choice of hardware bandwidth limits for the channel.  
4.2.6. Trigger  
Trigger characteristics are set using the settings in the control panel  
Trigger section. The trigger source setting can be either Internal (trigger on  
any one of the input channels) or External (trigger on an external input).  
The trigger source is indicated in the display by the name appearing under  
Source.  
In order to select the trigger source, click the drop-down source menu in  
the trigger Source section of the control panel and choose the desired one.  
Notice that, for most digitizer models, when external trigger is selected, the  
controls for the trigger termination, 50 Ω or 1 MΩ, are enabled in the  
Coupling section. Select the desired termination for the external trigger.  
The Coupling section also contains a selection for AC or DC coupling. AC coupling on the trigger sets the trigger  
circuit to an AC Low Frequency Reject mode.  
Use the Level section to set the desired trigger level. The relationship between coupling, trigger source, and trigger  
level was discussed in section 3.4.3 Trigger Level. The 10-bit digitizers have two adjustable trigger levels associated  
with the range of the window trigger.  
The trigger Delay setting positions the acquisition with respect to the trigger. Pre-trigger values are negative and are  
selectable in the range of [0 to – (10 * timebase)]. The value (10 * timebase) is equivalent to the time window or  
capture-time. 100% pre-trigger is achieved by setting the Delay to -capture-time. A 100% pre-trigger delay means  
the trigger point is at the right edge of the display and all of the data in the acquired waveform were captured just  
prior to the trigger event.  
Maximum Post trigger delay settings depend on the sample interval used. Post trigger delays are adjustable up to  
100 Mpoints.  
When a Slope box is shown next to the delay section, selecting the trigger slope is achieved by depressing either the  
positive or negative slope buttons. The 10-bit digitizers extend the functionality of the slope section to allow the  
choice of the window, HF divide, and Spike Stretcher trigger modes when the multiple mode slope button is selected.  
When the Window trigger is chosen buttons below it will allow the user to choose between trigger on entering or  
exiting the window trigger level range.  
The Adv button in this section gives access to the choice of External Trigger range and Bandwidth Limiter. It also  
allows the user to set the offset of the trigger out signal.  
4.2.7. Acquisition Mode  
The acquisition mode of the digitizers is selected using one of four  
acquisition mode buttons in the Acquisition section of the control panel.  
Available acquisition modes are Auto, Normal, Single and Stop.  
This Acquisition section of the control panel also indicates the number of  
waveforms acquired per second by the digitizer. The use of each of the acquisition modes is described below. In the  
User Manual: Family of 10-bit Digitizers  
Page 42 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
discussion, a valid trigger indicates a trigger signal meeting the trigger conditions at a time when the digitizer is  
armed and ready to acquire data. In Single and Normal modes the display will only be updated after each active  
digitizer has received a trigger.  
Stop will stop the acquisition and hold the latest complete acquisition on the display.  
Single mode is used in order to capture one event at the first valid trigger. It freezes the acquisition in the digitizer’s  
memory, and on the display, until the user requests another acquisition. After an acquisition is taken in Single mode,  
the digitizer will ignore subsequent trigger events until the Single button is pressed again or another acquisition mode  
is selected. Pressing the single button re-arms the trigger and captures one acquisition at the next valid trigger.  
Normal mode is used to continuously acquire waveforms into memory for valid trigger events. The display is  
updated with the new acquisitions in accordance with the Display Refresh Rate setting under the Options menu on  
the top line of the AcqirisLive Control window. Acquisitions will happen for each valid trigger until either single or  
Stop is pressed.  
If the trigger indicator at the lower left of the waveform display does not blink green, the digitizer is not receiving  
valid triggers. Check the trigger settings or select Auto mode in order to obtain a display of the waveform.  
Auto mode will acquire and display waveforms according to the trigger settings if a valid trigger is present within a  
timeout interval. If a valid trigger is not available within this interval, the digitizer generates its own trigger in order  
to digitize and display whatever signal is at the input at that time.  
If valid triggers are received at a high enough rate, Auto mode behavior differs little from normal mode behavior on  
the display. However if the trigger rate is too low or trigger settings are not appropriate for the characteristics of the  
signal, the input signal will not be stable on the display. This occurs because the internal auto-trigger generated by  
the digitizer is asynchronous to the input signal.  
Auto mode is often used to aid in setup when the input signal must be quickly characterized in order to determine  
proper trigger settings for Normal or Single mode acquisitions.  
4.2.8. Timebase and Memory  
The timebase and memory settings should be chosen together in order to  
optimize the desired sample rate for the acquisition.  
In oscilloscope mode the timebase settings range from 10 ns/div to 5 S/div.  
In Transient Recorder mode time windows of 100 ns to 20 Ks are available  
when using a card with the 2 Mpoint memory option.  
In order to change the timebase click on the white area in the timebase  
display window and select the desired timebase from the list.  
Alternatively, the Decrement / Increment timebase selection buttons  
may be used in order to change the timebase in steps with each click of the  
mouse. The sample rate and sample interval resulting from the combination of memory and timebase are indicated  
under the Sampling section of the control panel.  
The Memory setting shows the maximum number of points to be acquired. The actual number of points acquired  
could be less than the maximum number set if the acquisition time window selected fills less than the selected  
amount of memory at the full sample rate. Alternatively, in oscilloscope mode, if a slower timebase is selected, in  
order to ensure that the selected amount of memory will be filled before the acquisition is complete, the digitizer will  
reduce its sample rate as necessary. This action differs from the same case in Transient Recorder mode where  
preference is given to the selected sample interval.  
In Transient Recorder mode, the digitizer will continue to allocate more memory with increasing capture times. This  
is done in order to maintain the selected sample interval for as long a time window as is possible given the available  
memory. When the card memory is fully utilized, the sample rate will decrease with increasing capture times.  
4.2.9. Using digitizers of different models simultaneously  
AcqirisLive features a single timebase (i.e. the set of acquisition parameters pertaining to the time axis of the  
waveform, such as the sampling rate and the number of samples). It is therefore not possible to specify different  
timebase settings for different digitizers. However, it is possible to run each digitizer at its maximum available  
sampling rate performance in the following way:  
Select the Scope mode.  
In the channels list, select a channel from the fastest digitizer (that will make that digitizer the current  
digitizer).  
User Manual: Family of 10-bit Digitizers  
Page 43 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Make sure the maximum number of samples (MAX samples in Memory panel) is large enough.  
Select the desired time per division setting (timebase panel): the fastest sampling rate for the current  
digitizer, compatible with the maximum number of samples, will be selected and applied to all digitizers  
in the system. Any digitizer not capable to satisfy that sampling rate will run at its closest setting.  
Finally, note that the sampling rate menu (visible in Transient Recorder mode) contains the sampling rate capability  
of all digitizers currently in use. You must turn off the channels of 'slower' digitizers in order to view the highest  
sampling rates of the 'faster' digitizers.  
4.2.10. Segmented Memory  
Segmented memory enables the sequential acquisition mode. This feature is particularly useful when capturing  
impulse-response type waveforms in a variety of applications. Segmented memory is available in all acquisition  
modes. However, it is generally used in either Normal or Single mode. Waveform display occurs only after all  
segments are filled. Segments can be displayed sequentially on the display or they can be overlaid on top of each  
other. To select an overlaid display, select the item Overlay Segments under the Options menu. Segmented data can  
be archived to any storage device on the computer in a similar way to non-segmented acquisitions. If the acquisition  
needs to be stopped before all segments are filled, press the Stop button.  
In order to set up the sequential acquisition mode, simply select more than 1 segment on the segment  
selection control under the Memory section of the control window and start the acquisition.  
4.2.11. Display Features, Zoom and Persistence  
A waveform zoom feature is available under the View menu of the  
display window. The screen image below shows the normal display  
and the zoom window that appears when zoom is selected.  
The zoom scale and center position are modified using the horizontal  
and vertical controls shown on the zoom display window. To the right  
side of the zoom window click on the “Horizontal” button, then  
modify the zoom expansion factor by clicking the “In” or “Out”  
buttons.  
The vertical blue cursor, or box, (depending on the zoom factor),  
shown on the main display window indicates the center point and  
range of the zoom. The vertical box or line can be dragged right or left  
with the mouse in order to re-position the center point of the  
waveform displayed in the zoom display window. Alternatively, the  
right and left arrows at the bottom of the zoom display window will  
reposition the horizontal center of the zoom.  
In order to expand the zoom region vertically, select the “Vertical”  
button on the zoom display and click the “In” / “Out” buttons. Alternatively, the up / down arrows on the scale  
indicator just below the vertical button can be used. When the vertical scale is expanded, the top and bottom lines  
forming the blue box in the main display window show the extent of the vertical zoom region. The vertical position  
of the box, and thus the data region shown on the zoom display, may be changed by moving the box vertically with  
the mouse or by clicking on the up and down arrows at the left of the zoom display window.  
When expanding a waveform to the point where individual data points are seen, it may be desirable to use the  
“Connect data points” option under the Options menu in AcqirisLive. When this option is selected, larger data points  
are drawn on the display and a line connects each point.  
The main display window contains a drop down selector that controls the display type for both the main display and  
the zoom window. There are two modes available, the Live Display, or the Persistence Display. When persistence  
display is selected, all points drawn to the display windows remain on while new points are added from subsequent  
acquisitions. This mode is particularly suited for examining the jitter on an edge of a periodic signal as shown in the  
Us
Page 44 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
image below.  
To clear the display in persistence display mode, press the “Clear” button that appears on the main display window.  
4.3.  
Top Line Menu of AcqirisLive  
The top line menu contains functions offering additional utility to  
the AcqirisLive application. Provisions are included for storing  
instrument setups, waveforms, converting the format of stored  
data files, identifying models and versions and in the Options  
Menu, calibrating the digitizer, extended control of event acquisition and controlling the display properties. These  
possibilities are discussed in the next few sections.  
4.4.  
Setup  
The Setup menu allows you to store and recover instrument configurations. When you store a configuration the  
parameters used to configure each of the channels of each module together with the timebase and trigger  
configurations will be saved in a file. You will be asked to choose the destination directory for the file and be  
allowed to give it an appropriate name. When you choose to recover a file you will be asked to give the name or  
navigate to it using the browser functionality. If you are using many modules the settings in the configuration file  
will be associated preferentially with the digitizer that has that serial number or then with the digitizer in the same  
location.  
Only the parameters of the modules are stored in the configuration file. This means that parameters of the  
AcqirisLive program, such as channel activation and display mode choices are not included.  
4.5.  
Store, Autostore and Conversion of  
Waveforms  
The store menu gives you two mechanisms to store waveforms.  
1. The Store Waveform option allows you to store the last  
acquired waveform for the current channel. To store the  
last acquired waveform for each of the active channels,  
you must sequentially select each one to make it the  
current channel and use the Store Waveform menu  
command. The title of the Store Waveform dialog  
indicates the name of the channel that will be stored.  
The data is stored in ASCII format, as Raw ADC  
values or Amplitudes (in Volts).  
2. The Autostore option allows you to store consecutively  
acquired waveforms from all of the active channels into  
files. Configuration of Autostore is achieved by  
selecting the Autostore item under the Store menu.  
Autostore brings up the dialog box shown here. Three  
formats are available for stored waveforms, Binary,  
Raw ADC values, and Amplitude values (in Volts). The destination folder can be selected or created on any  
system drive using the Browse button. Waveforms will be stored in a new folder under the destination  
folder. The new folder will be named AcqirisRun_{date}_{time}. Each waveform will be stored in a file  
with a name {channel identifier}_{i}.{suffix}, where i goes from 1 to N and the suffix indicates the data  
type. If the data is stored in Binary form, the suffix will be .data and an additional ASCII format .awh file is  
written which contains necessary details about each of the other files. Otherwise, the suffix will be .awd.  
The number of generations of waveforms to be stored with this function is selectable from 1 to N. N is a  
maximum number dependent on the number of samples taken in each waveform. There is also a user  
selectable limit on the total amount of data allowed in any single Autostore acquisition series.  
When using segmented memory with Autostore, a waveform is considered to be a complete set of acquired  
segments. When storing Amplitudes, it is possible to write each segment as a separate column of data in the  
waveform file. This is achieved by clicking the check box option Segments as columns in the Autostore  
settings window. Formatting with segments in columns, in Amplitude format, is limited to 200 segments.  
The Binary file format selection provides the fastest storage performance and generates the smallest files on  
the hard drive. Binary files can be subsequently converted to either Raw ADC values or to Amplitude  
values using the Data Format Conversion utility provided with AcqirisLive.  
User Manual: Family of 10-bit Digitizers  
Page 45 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Selecting the “Write files at end option” acquires all waveforms into memory first, before converting to the  
selected format and writing the data to disk. This option allows for faster data capture times. Selecting the  
“Show waveforms” allows the user to see the waveforms before writing them to disk.  
When all parameters are selected, click OK, and then select Normal, Auto or Single acquisition mode.  
When the waveforms are recorded a progress bar will appear to show the status of the acquisition and of  
writing the data to disk. The Autostore function can be aborted from the progress bar.  
Raw ADC values are the raw 16-bit ADC levels in ASCII format. The channel’s gain and offset are needed in order  
to obtain the signal level in Volts. The required information to convert the Raw ADC values to voltage is included in  
the header information of each Raw ADC ASCII file or in the .awh file accompanying the binary files. In order to  
convert the Raw ADC values into Volts the following formula should be used.  
V = vGain * data – vOffset  
The Amplitudes [V] selection stores the waveforms as voltage samples in ASCII format. This data has already been  
adjusted for gain and offset and can be directly read out as a voltage. However, it requires roughly twice as much  
space as Raw ADC value data.  
For the sequential acquisition mode, the segment time stamps are also stored when waveforms are saved using the  
Autostore and Save Waveform functions.  
Data stored in binary format with Autostore must be converted before it  
can be imported into a data processing package or a spreadsheet (e.g.  
Excel). The Data Format Conversion option is found under the Store menu.  
Select Data Format Conversion in order to bring up the following window.  
The acquisition must be stopped in order to access this menu.  
To convert the binary data, simply select the source folder where the data  
is located, select the output format and click OK.  
All binary data files found in the source directory will be converted to the  
desired format. If you are writing amplitudes with multiple segments,  
formatting with segments in columns is limited to 200 segments.  
4.6.  
Calibrate  
The Calibrate item in the Options menu leads to a sub-menu where a particular type of calibration can be invoked. In  
order to use this item the acquisition must be stopped; otherwise the Calibrate option is not selectable. The  
possibilities are  
ƒ
Calibrate instruments - to perform a self-calibration on the digitizers, calibrating the gain, offset, trigger  
level and the timing. Performing a calibration compensates for any drift in these properties, typically due to  
component temperature variations.  
ƒ
ƒ
Calibrate current combine configuration – to perform a calibration only at the current configuration in each  
instrument.  
Calibrate External Clock Timing – to do a calibration of the internal TDC in the Continuous External Clock  
mode.  
4.7.  
Combine channels  
The Combine Channels function in the Options menu give the 10-bit digitizer user the control of how many  
converters will be used for each channel. This allows the memory size and the sampling rate to be increased in  
situations where not all of the input channels need to be acquired. Normally digitizer modules work with one  
converter/channel. If two or more converters/channel are desired, the identities of the channels to be used have to be  
chosen with the Active Channel control.  
4.8.  
External 10 MHz Reference and External Clock  
Either one of the External 10 MHz Reference or External Clock options can be selected in the Options Menu. The  
external input threshold option, available in the options menu, must also be set appropriately. The 10 MHz Reference  
can be activated at the I/O B output.  
The External Clock function is available in AcqirisLive only when using the control panel in Transient Recorder  
mode. The input frequency should be keyed into the clock frequency selector (outlined in red) on the control panel.  
Then the appropriate sampling frequency must be selected in order for the proper timebase to be displayed on the  
User Manual: Family of 10-bit Digitizers  
Page 46 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
waveform display window. The time window displayed can be manipulated by allocating more or less memory to the  
acquisition.  
The External 10 MHz Reference in AcqirisLive is designed to work either in Oscilloscope mode or in Transient  
Recorder mode. The timebase and sample rate for waveform capture is fully selectable when using the external  
reference, just as it is when using the digitizer’s own internal reference clock.  
4.9.  
Additional Waveform Information  
The Options menu also includes possibilities to obtain additional information about the waveform. In both cases  
described below the display will be updated after each acquisition.  
The “Show Segment Timestamps” option shows a display like the one below:  
It gives you the information needed to better understand the trigger times of the  
individual segments of an acquisition. By definition the first trigger of an  
acquisition occurs at t1=0.0. There are two columns for each digitizer. For the n’th  
row, the first column contains the difference tn – tn-1 and the second column  
contains tn. The top line contains the average value of the tn in the left column and  
the standard deviation of the tn in the right one. The times are in ps.  
The “Show Parameters” option activates a display of pulse parameters for each  
active channel. Inactive channels will have parameter values of 0.0. The parameters  
will be calculated in a model in which the distribution of measured voltages has  
two distinct peaks. If this condition is not satisfied the results will be unreliable.  
4.10. Display Features  
Display features in the Options menu include items for  
turning the waveform display off (by de-selecting “Display Waveforms”),  
overlaying segments of a sequence,  
connecting displayed data points,  
modifying the display refresh rate,  
displaying the internal temperature of the currently selected module  
4.11. Command Line Switches  
When starting the AcqirisLive application, the executable's name in the command line can be followed by one or  
more switches to change the running conditions.  
Syntax  
The syntax of the command line is the following:  
path\AcqirisLive.exe options  
where options are chosen from the following list  
[-a][-b][-B][-c][-d][-h][-i][-l][-m][-n][-p][-s][t][-v][-x]  
Available Switches  
List of available switches:  
-a  
-b  
-B  
-c  
-d  
-h  
-i  
start in Averager mode  
Buffer size for large waveforms  
Number of bank for SMAR mode  
no initial calibration  
use DMA for data transfers (default)  
high priority process (default)  
use acquisition interrupts  
User Manual: Family of 10-bit Digitizers  
Page 47 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
-l  
low priority process  
-m  
-n  
-p  
-r  
-s  
medium priority process  
no DMA for data transfers  
Use acquisition polling  
start in Transient Recorder (digitizer) mode  
simulation mode  
-t  
temperature update off  
-v  
-x  
show acquisition state  
disable automatic multi-instruments  
Operation Mode  
Acqiris averagers can be operated in Averager Mode or in standard Digitizer Mode.  
AcqirisLive operates in either mode, but since these are mutually exclusive, only instruments that support the chosen  
mode are accessible in any given mode. E.g. only averagers are accessible in Averager Mode, while both digitizers  
and averagers are accessible in Digitizer Mode.  
The operation mode can be switched within the application. By default, AcqirisLive starts in Digitizer Mode.  
Buffer size  
Upon startup of the application, the available physical memory on the computer is checked. The total data buffer size  
is then limited to the available memory minus 10% of the total memory. It is also possible to limit the total data  
buffer size to a fixed value with the -bn option (where n is the limit in MB, e.g. -b200 will limit the buffer to 200  
MB). The data buffer is then shared among the active channels.  
When the total acquisition memory (sum over all active channels) exceeds the data buffer size limit, readout of the  
data will be partial. The special operation modes such as Segments Average, Average, Digital Filter and FFT are not  
supported with partial data readout.  
Partial readout is controlled with the Very Long Memory panel (menu Options ->Very Long Memory...). The first  
segment and the first sample to read out can be specified. Note that if the buffer is large enough for at least 1 full  
segment, only full segments will be read (i.e. first sample control is ignored).  
When saving data to file, you will be prompted for a maximum number of segments per file.  
Banks for SMAR  
Some of the 10-bit Family modules are capable of acquiring data and being read out simultaneously. The -Bn option  
allows the internal memory of the digitizer to be divided into n banks. To observe the improved acquisition trigger  
rate achievable in this mode it is suggested to set up the desired conditions and then turn the event display off. The  
trigger out signal can be used to monitor the trigger rate; an interval width trigger can be used to check if triggers are  
missing.  
Calibration  
Before operation, the instruments must undergo an internal calibration. As this calibration can only proceed  
sequentially for each channel, the initialization time of AcqirisLive can become significant for systems with many  
channels. Therefore, it can sometimes be desirable to be able to start the application quickly by skipping the  
calibration (which must then be done "manually" before collecting correct data).  
DMA Data Transfers  
By default, data transfers on the PCI bus between the digitizer memory and the PC are done in DMA (Direct  
Memory Access) mode, which allows the fastest transfer speeds (up to about 130 MB/s peak, 100 MB/s sustained  
acquisition data rate). These very high data rates are not well supported by all hardware configurations (PC  
motherboard), and system hangs due to endless retries on the PCI bus have been observed on some machines. Such  
problems disappear when switching DMA mode off (data transfer rate is then about 2.8 MB/s).  
Please note that switching DMA data transfer mode off precludes the use of acquisition interrupts. The –i switch will  
therefore be ignored when –n is used.  
Process Priority  
User Manual: Family of 10-bit Digitizers  
Page 48 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
When the program checks for an event from the operating system (e.g. to react to user input), it can put itself to sleep  
for a specified period of time. This gives other applications more processor time, but limits the performance of  
AcqirisLive in terms of maximum acquisition rate. The default is not to go to sleep, but this can be overridden to  
allow better performance for other applications. Note that this sleep policy is only in effect while the acquisition is  
running. When the acquisition stops, the sleep policy is always reset to the longest sleep period.  
Acquisition Interrupts  
When performing acquisitions, there are 2 possible ways to wait for the acquisition to complete: polling or using a  
hardware interrupt.  
-p Polling: AcqirisLive runs as a single thread process. After an acquisition is started, the instruments are  
regularly polled to check when the acquisition is complete. The user interface is still alive, but the latency  
for reading the data after the acquisition may vary.  
-i Hardware interrupt: In this mode, AcqirisLive is run as a multi-thread process. After starting an  
acquisition, a new thread is created to wait for the hardware interrupt indicating that the acquisition  
completed. This thread is put to sleep, and wakes up upon reception of the interrupt from the instrument, or  
after a 2 seconds timeout.  
Please note that switching the DMA data transfer mode off precludes the use of acquisition interrupts. The –i switch  
will therefore be ignored when –n is used.  
Transient Recorder mode  
The application will start in the Transient Recorder mode as opposed to the oscilloscope mode. In this case, if a file  
name of an AqConfig setup file is given on the command line it will be loaded.  
Simulation Mode  
Simulation mode allows you to run AcqirisLive without having any digitizer installed in the PC.  
Temperature Update  
The internally measured temperature of the currently selected module is normally displayed. Using this switch turns  
this feature off.  
Acquisition State  
This switch turns on the display of the current acquisition state (Stopped, Ready, Started, Reading, Data ready). This  
is helpful when running with time-consuming settings (e.g. many segments overlaid, many channels, large records,  
etc.), but will slow down the overall performance.  
Multi-instruments  
By default, AcqirisLive tries to detect instruments connected through ASBus, and will initialize any connected  
digitizers as a multi-instrument. This behavior can be inhibited by the -x switch, in which case any ASBus connector  
will be ignored.  
User Manual: Family of 10-bit Digitizers  
Page 49 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5.  
Running the GeoMapper Application  
5.1.  
Who needs a Geographical Map of Modules  
The GeoMapper application gives the user the possibility of telling the Acqiris driver crucial details of the physical  
position of modules when this information cannot be deduced directly. It allows the driver to correctly adjust trigger  
and clock delays between the different modules of an ASBus2 multi-Instrument and arrange the modules in an  
intuitive order.  
The GeoMapper application has been implemented as a Wizard with the goal of learning the physical position of  
each module of a multi-bus system. The driver already knows the logical Bus and logical Slot numbers of each  
Acqiris unit present. However, because of varying conventions for the compactPCI slot arrangement on a bus, this is  
not sufficient to determine the physical position. Furthermore, with the aid of the map, the software can ensure that  
the modules of an Auto-defined multi-Instrument will be ordered to correspond to increasing physical slot number.  
NOTE: The unusual slot numbering convention of the CC121 is documented in the CC121 CompactPCI Crate User  
Manual. The CC10X crates have a simpler slot numbering convention that the Acqiris driver already takes into  
account. CompactPCI Crates from other manufacturers represent an unknown element.  
NOTE: The GeoMapper is of use for some ASBus2 systems in CC121 crates and, perhaps, crates from other  
manufacturers. CC121 crate configurations with both an Acqiris acquisition module in one of the last 7 slots and  
with the PC running under Windows 2000, NT, or XP will not need the GeoMapper. Users who do not need  
GeoMapper do not need to read this chapter any further.  
5.2.  
When should the GeoMapper Application be used  
The application needs to be used whenever the bus or slot configuration of the system changes in a fundamental way.  
The most obvious trouble symptom is that an expected multi-Instrument was not found. Alternatively the driver can  
issue error messages that the AcqirisLive application and other programs will see. Among them are  
Modules are not contiguous in ASBus2 combination  
ACQIRIS_ERROR_MODULES_NOT_CONTIGUOUS  
ASBus2 Candidate Modules not on same bus  
'AcqGeo.map' file is invalid or corrupted  
ACQIRIS_ERROR_MODULES_NOT_ON_SAME_BUS  
ACQIRIS_ERROR_INVALID_GEOMAP_FILE  
In all of these cases the GeoMapper application should be run to ensure that the map file is up to date.  
If desired the GeoMapper application will generate a default configuration in which it allows the driver to make all of  
the choices. This is done by specifying "None (Disable GeoMap)" as the desired configuration in "Step1: Crate  
count" of the application.  
5.3.  
How to run GeoMapper  
After software installation you will find GeoMapper in the AcqirisApp directory of the installation directory. After  
being started the program will learn about the Acqiris devices present and then show a screen like the one below:  
The choice of the number of crates ranges from None up to the number of bus segments that the driver has found  
containing Acqiris modules. If you choose None and save the resulting AqGeo.map file the system will behave in its  
natural way without any additional user provided information. If you want to give the driver information on the  
physical position of modules you should select the appropriate number of crates. If the number of crates does not  
correspond to your expectations you may have hardware problems (unpowered, broken, or badly cabled crates or  
modules).  
The next step is to associate each bus segment with a physical crate. The menu to do this is shown below:  
User Manual: Family of 10-bit Digitizers  
Page 50 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Selecting a row will cause all of the LED's of modules in that Bus to be turned on RED. This should make it easy to  
identify which crate the modules are in.  
The third step is to specify the number of slots in each crate. This is shown below:  
The 4th step consists of arranging the modules of each crate into the correct order. The menu shown below is one in  
which all of this work has been done for the modules of a fully populated CC121. Selecting a position will cause a  
module to have its red LED turned on. Then the Up and Down buttons will allow you to move the module in  
question to the desired "Position in crate".  
The final step is to actuate the Save button so that the final configuration will be memorized and made available to  
the driver.  
User Manual: Family of 10-bit Digitizers  
Page 51 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
The GeoMapper application will always work with the standard AqGeo.map file. If you need to have several such  
files corresponding to different configurations it is up to you to either rename the files or transfer them to other  
directories in an appropriate manner.  
User Manual: Family of 10-bit Digitizers  
Page 52 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
6.  
Appendix A: ASBus2 for CompactPCI/PXI Digitizers  
Specification and User Instructions  
User Manual: Family of 10-bit Digitizers  
Page 53 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
7.  
Appendix B: XA001 Battery Backup for CompactPCI Digitizers  
Specification and Assembly Instructions  
User Manual: Family of 10-bit Digitizers  
Page 54 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
8.  
Appendix C: XA100 BNC Input Overvoltage Protection  
Specifications and User Instructions  
User Manual: Family of 10-bit Digitizers  
Page 55 of 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 

3Com Modem MIC 3365 User Manual
3Com Tablet Accessory 86 0621 000 User Manual
Abocom Network Router AR3500 User Manual
Adcom Stereo Amplifier GFA 555se User Manual
Aiphone Network Card PD 1 User Manual
Altinex Water Dispenser PNP515 User Manual
American Standard Guitar 192500 User Manual
Audiovox Automobile Electronics ACA200 User Manual
Beko Flat Panel Television 19WLP530HID User Manual
Eureka Tents Tent Scenic Pass XT User Manual