Agilent Technologies Electronic Accessory U1251A User Manual |
Agilent U1251A and
U1252A
Handheld Digital
Multimeter
User’s and Service Guide
Agilent Technologies
Download from Www.Somanuals.com. All Manuals Search And Download.
Safety Symbols
The following symbols on the instrument and in the documentation indicate precautions which must be taken to
maintain safe operation of the instrument.
CAT IV
600 V
CAT III
1000 V
Category III 1000 V overvoltage protection
Category IV 600 V overvoltage protection
Earth (ground) terminal
Equipment protected throughout by
double insulation or reinforced
insulation
Caution, risk of danger (refer to this manual
for specific Warning or Caution information)
Caution, risk of electric shock
Agilent U1251A/U1252A User’s and Service Guide
III
Download from Www.Somanuals.com. All Manuals Search And Download.
Safety Information
This meter is safety-certified in compliance with EN/IEC 61010-1:2001, UL 61010-1 Second Edition and CAN/CSA
22.2 61010-1 Second Edition, CAT III 1000 V/ CAT IV 600 V Overvoltage Protection, Pollution Degree II. Use with
standard or compatible test probes.
General Safety Information
The following general safety precautions must be observed during all phases of operation, service, and repair of
this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual
violates safety standards of design, manufacture, and intended use of the instrument. Agilent Technologies
assumes no liability for the customer’s failure to comply with these requirements.
IV
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
• When working above 70 VDC, 33 VAC RMS or 46.7 V peak, exercise caution – such
range pose a shock hazard.
WARNING
• Do not measure more than the rated voltage (as marked on the meter) between
terminals, or between terminal and earth ground.
• Double-check the meter’s operation by measuring a known voltage.
• For current measurement, turn off circuit power before connecting the meter to the
circuit. Always place the meter in series with the circuit.
• When connecting probes, always connect the common test probe first. When
disconnecting probes, always disconnect the live test probe first.
• Detach test probes from the meter before you open the battery cover.
• Do not use the meter with the battery cover or part of the cover removed or loose.
• Replace the battery as soon as the low battery indicator
flashes on screen. This is
to avoid false readings, which may lead to possible electric shock or personal injury.
• Do not operate the product in an explosive atmosphere or in the presence of
flammable gases or fumes.
• Inspect the case for cracks or missing plastic. Pay extra attention to the insulation
surrounding the connectors. Do not use the meter if it is damaged.
• Inspect the test probes for damaged insulation or exposed metal, and check for
continuity. Do not use the test probe if it is damaged.
• Do not use any other AC charger adaptor apart from the one certified by Agilent with
this product.
• Do not use repaired fuses or short-circuited fuse-holders. For continued protection
against fire, replace the line fuses only with fuses of the same voltage and current
rating and recommended type.
• Do not service or perform adjustments alone. Under certain condition, hazardous
voltages may exist, even with the equipment switched off. To avoid dangerous electric
shock, service personnel must not attempt internal service or adjustment unless
another person, capable of rendering resuscitation or first aid, is present.
• Do not substitute parts or modify equipment to avoid the danger of introducing
additional hazards. Return the product to Agilent Technologies Sales and Service
Office for service and repair to ensure the safety features are maintained.
• Do not operate damaged equipment as the safety protection features built into this
product may have been impaired, either through physical damage, excessive
moisture, or any other reason. Remove power and do not use the product until safe
operation can be verified by service-trained personnel. If necessary, return the
product to Agilent Technologies Sales and Service Office for service and repair to
ensure the safety features are maintained.
Agilent U1251A/U1252A User’s and Service Guide
V
Download from Www.Somanuals.com. All Manuals Search And Download.
• Turn off the circuit power and discharge all high-voltage capacitors in the circuit before
you perform resistance, continuity, diodes, or capacitance tests.
• Use the correct terminals, function, and range for your measurements.
• Never measure voltage when current measurement is selected.
• Use only recommended rechargeable battery. Ensure proper insertion of battery in the
meter, and follow the correct polarity.
CAUTION
• Disconnect test leads from all the terminals during battery charging.
VI
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Regulatory Markings
The C-tick mark is a registered
trademark of the Spectrum
The CE mark is a registered trademark
of the European Community. This CE
mark shows that the product complies
with all the relevant European Legal
Directives.
Management Agency of Australia. This
signifies compliance with
the Australia EMC Framework
regulations under the terms of the
Radio Communication Act of 1992.
This instrument complies with the
WEEE Directive (2002/96/EC) marking
requirement. This affixed product label
indicates that you must not discard
this electrical/electronic product in
domestic household waste.
ICES/NMB-001 indicates that this ISM
device complies with the Canadian
ICES-001.
Cet appareil ISM est confomre a la
norme NMB-001 du Canada.
The CSA mark is a registered
trademark of the Canadian Standards
Association.
Agilent U1251A/U1252A User’s and Service Guide
VII
Download from Www.Somanuals.com. All Manuals Search And Download.
Waste Electrical and Electronic Equipment (WEEE) Directive
2002/96/EC
This instrument complies with the WEEE Directive (2002/96/EC) marking
requirement. This affixed product label indicates that you must not discard
this electrical/electronic product in domestic household waste.
Product Category:
With reference to the equipment types in the WEEE directive Annex 1, this
instrument is classified as a “Monitoring and Control Instrument” product.
The affixed product label is as shown below.
Do not dispose in domestic household waste
To return this unwanted instrument, contact your nearest Agilent
Technologies, or visit: www.agilent.com/environment/product for more
information.
Agilent Technologies, through Rechargeable Battery Recycling Corporation (RBRC), offers free and convenient
battery recycling options in the U.S. and Canada. Contact RBRC at 877-2-RECYCLE (877.273.2925) or online at:
http://www.call2recycle.org/ for the nearest recycling location.
VIII
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
In This Guide…
1
and rear panel.
2
3
4
5
the U1251A and U1252A handheld digital multimeter.
available for the U1251A and U1252A digital multimeter.
Changing The Default Setting
This chapter shows you how to change the default factory settings of the
U1251A and U1252A and other available setting options
This chapter will go through how to troubleshoot the handheld digital
multimeter if any problems arise.
6
7
This chapter contains the performance test and the adjustment procedure.
Specifications
This chapter lists the product characteristics, specification assumptions
and the specifications of the U1251A and U1252A digital multimeters.
Agilent U1251A/U1252A User’s and Service Guide
IX
Download from Www.Somanuals.com. All Manuals Search And Download.
Declaration of Conformity (DoC)
The Declaration of Conformity (DoC) for this instrument is available on the
Web site. You can search the DoC by its product model or description.
http://regulations.corporate.agilent.com/DoC/search.htm
NOTE
If you are unable to search for the respective DoC, please contact your
local Agilent representative.
X
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Contents
1
2
Adjusting the tilt-stand
The display at a glance 11
Selecting display with the Dual button 18
Selection display with the Shift button 20
2
A (ampere) measurement 31
Frequency Counter 32
Measuring Resistance, Conductance and Testing Continuity 34
Testing Diodes 38
Measuring Capacitance 41
Measuring Temperature 43
Alerts and Warning During Measurement 47
Abbreviated Book Title Variable (Edit this)
XIII
Download from Www.Somanuals.com. All Manuals Search And Download.
III
Overload alert 47
Input warning 47
3
Functions and Features
Dynamic Recording 50
Data Hold (Trigger Hold) 52
Refresh Hold 53
Interval logging 63
4
Changing The Default Setting
Setting Minimum Frequency Measurement 81
Setting Temperature Unit 82
Setting Auto Power Saving Mode 84
Setting Percentage (%) Scale Readout 86
XIV
Abbreviated Book Title Variable (Edit this)
Download from Www.Somanuals.com. All Manuals Search And Download.
III
Setting Baud Rate 89
Setting Parity Check 90
Setting Data Bit 91
Returning to Default Factory Settings 94
5
Storage considerations 98
Charging the battery 98
Replacing the fuse 106
Troubleshooting 108
Replaceable Parts 109
To order replaceable parts 109
6
Calibration Overview 112
Closed-case electronic calibration 112
Adjustment is recommended 113
Recommended Test Equipment 114
Basic Operating Test 115
Backlit test 115
Abbreviated Book Title Variable (Edit this)
XV
Download from Www.Somanuals.com. All Manuals Search And Download.
III
Valid adjustment input values 134
Finishing the adjustment 142
To read the calibration count 142
7
Capacitance Specifications 156
Temperature Specifications 156
Frequency Specifications 157
Duty Cycle and Pulse Width Specifications 157
Frequency Sensitivity Specifications 158
XVI
Abbreviated Book Title Variable (Edit this)
Download from Www.Somanuals.com. All Manuals Search And Download.
III
Square Wave Output for U1252A 161
Operating Specifications 162
Display update rate (approximate) 162
Input impedance 163
Abbreviated Book Title Variable (Edit this)
XVII
Download from Www.Somanuals.com. All Manuals Search And Download.
III
XVIII
Abbreviated Book Title Variable (Edit this)
Download from Www.Somanuals.com. All Manuals Search And Download.
List of Figures
4
4
5
6
7
8
9
Figure 2-4 Measuring scale of 4-20 mA 30
test. 35
Figure 2-9 Conductance measurement 37
Figure 2-11 Measuring reverse bias of diode 40
Figure 3-3 Refresh hold mode operation 54
Figure 3-4 Null (relative) mode operation 56
Figure 3-5 dBm/dBV display mode operation 58
Figure 3-6 1 ms peak hold mode operation 60
Figure 3-7 Hand (Manual) logging mode operation 62
Abbreviated Book Title Variable (Edit this)
IX
Download from Www.Somanuals.com. All Manuals Search And Download.
Figure 3-8 Full Log 62
Figure 3-9 Interval (Automatic) logging mode operation 64
Figure 3-10 Log review mode operation 66
Figure 3-11 Frequency adjustment for square wave output 68
Figure 3-12 Pulse width adjustment for square wave output 69
Figure 3-13 Duty cycle adjustment for Square Wave 70
Figure 3-14 Cable connection for remote communication 71
Figure 4-1 Data hold/Refersh hold setup 77
Figure 4-2 Data logging setup 78
Figure 4-3 Thermocouple type setup 79
Figure 4-4 Reference impedance for dBm measurement setup 80
Figure 4-5 Minimum frequency setup 81
Figure 4-6 Temperature unit setup 83
Figure 4-7 Auto power saving setup 85
Figure 4-8
% scale readout setup 86
Figure 4-9 Beep frequency setup 87
Figure 4-10 Backlit timer setup 88
Figure 4-11 Baud rate setup remote control 89
Figure 4-12 Parity check setup 90
Figure 4-13 Data bit setup for remote control 91
Figure 4-14 Echo mode setup for remote control 92
Figure 4-15 Print mode setup for remote control 93
Figure 4-16 Reset setup 94
Figure 5-1 9 V rectangular battery 97
Figure 5-2 Battery capacity display as trickle 100
Figure 5-3 Self-test 100
Figure 5-4 Charging mode 102
Figure 5-5 Charge end and trickle state 103
Figure 5-6 Battery charging procedure 104
Figure 5-7 Fuse replacement 107
Figure 6-1 LCD display 115
Figure 6-2 Input warning 116
Figure 6-3 Charge terminal alert 117
X
Abbreviated Book Title Variable (Edit this)
Download from Www.Somanuals.com. All Manuals Search And Download.
II
List of Tables
8
Table 1-6 Analog bar range and counts 15
9
functions 22
Table 2-1 Numerical steps descriptions 24
Table 6-1 Recommended Test Equipment 114
Table 6-3 Valid adjustment input values 134
Table 7-1 DC Accuracy (% of reading + Number of Least Signifi-
Table 7-2 U1251A accuracy specifications (% of reading + num-
Table 7-3 U1251A accuracy specifications (% of reading + num-
Table 7-4 U1252A accuracy specifications (% of reading + num-
ber of LSD) for true RMS AC voltage 154
Table 7-5 U1252A accuracy specifications (% of reading + num-
ber of LSD) for true RMS AC current 154
Table 7-6 U1252A true RMS ac+dc voltage specifications 155
Abbreviated Book Title Variable (Edit this)
XI
Download from Www.Somanuals.com. All Manuals Search And Download.
II
Table 7-7 U1252A true RMS ac+dc current specifications 155
Table 7-8 Capacitance specifications 156
Table 7-9 Temperature specifications 156
Table 7-10 Frequency specifications 157
Table 7-11 Duty cycle and pulse width specifications 157
Table 7-12 Frequency sensitivity and trigger level specifications for
Table 7-13 Frequency sensitivity specifications for current
measurements 159
Table 7-14 Peak hold specifications for dc voltage and current
measurements 159
Table 7-15 Frequency counter (divide by 1) specifications 160
Table 7-16 Frequency counter (divide by 100 [4])
specifications 160
Table 7-17 Square wave output specifications 161
Table 7-18 Display update rate (approximate) 162
Table 7-19 Input Impedance 163
XII
Abbreviated Book Title Variable (Edit this)
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent U1251A and U1252A Handheld Digital Multimeter
User’s and Service Guide
1
2
Check the shipment
Adjusting the tilt-stand
The front panel at a glance
The rear panel at a glance
The rotary switch at a glance
The keypad at a glance
The display at a glance 11
Selecting display with the Hz button 16
Selecting display with the Dual button 18
Selection display with the Shift button 20
The terminals at a glance 22
This chapter contains information on the Agilent U1251A
and U1252A handheld multimeter front panel, rotary switch,
keypad, display, terminals and rear panel.
Agilent Technologies
1
Download from Www.Somanuals.com. All Manuals Search And Download.
1
Getting Started Tutorial
Introducing the U12521A/U1252A Handheld Digital Multimeter
Key features of this digital multimeter:
• DC, AC, and AC + DC (U1252A only) voltage and current
measurements.
• True-RMS measurement for both AC voltage and current
• Rechargeable Ni-MH battery with built-in charging
capability (U1252A only)
• Ambient temperature on second display
• Battery capacity indicator
• Blue LED backlight
• Resistance measurement of up to 50 MΩ (for U1251A)
and 500 MΩ (for U1252A)
• Conductance measurement from 0.01 nS (100 GΩ) ~ 50 nS
• Capacitance measurement of up to 100 mF
• Frequency counter up of to 20 MHz (U1252A only)
•
The % scale readout for 4–20 mA or 0–20 mA measurement
• dBm with selectable reference impedance
• 1 ms Peak Hold to catch inrush voltage and current easily
• Temperature test with selectable 0 °C compensation
(without ambient temperature compensation).
• K-type (for U1251A) and J/K-types temperature
measurement (for U1252A)
• Frequency, duty cycle, and pulse width measurements
• Dynamic Recording for min, max, and average readings
• Data Hold with manual or auto trigger and Null mode
• Diode and audible continuity tests
• Square wave generator with selectable frequency, pulse
width and duty cycle (U1252A only)
• Agilent GUI Application Software (IR-USB cable sold
separately)
• Closed case calibration
2
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Getting Started Tutorial
1
Check the shipment
Verify that you have received the following items with your
multimeter:
• 9 V alkaline battery (for U1251A only)
• Soft carrying case
• 4 mm probes
• Test leads
• Alligator clips
• SMT grabbers
• Fine tips test probes
• Mini grabber (black only)
• Rechargeable 7.2 V battery (for U1252A only)
• Power cord & AC adapter (for U1252A only)
• Quick Start Guide
• CD containing the User’s Guide, application software and
instrument drivers
• Certificate of Calibration
Contact your nearest Agilent Sales Office if any of the above
are missing.
Inspect the shipping containing for damage. Signs of damage
may include a dented or torn shipping container or
cushioning material that indicates signs of unusual stress or
compacting. Save the packaging material in case the
multimeter needs to be returned.
(5989-7340EN) for the full and latest list of handheld
accessories available.
Agilent U1251A/U1252A User’s and Service Guide
3
Download from Www.Somanuals.com. All Manuals Search And Download.
1
Getting Started Tutorial
Adjusting the tilt-stand
To adjust the meter to a 60° standing position, pull the
tilt-stand outwards to its maximum reach.
IR-USB cable
To PC (host)
Tilt-stand at 60 °
Figure 1-1 Tilt-stand at 60°
To adjust the meter to a 30° standing position, bend the tip
of the stand so that it is parallel to the ground before
pulling the stand outwards to its maximum reach.
IR-USB cable
To PC (host)
Tilt-stand at 30 °
Figure 1-2 Tilt-stand at 30°
4
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Getting Started Tutorial
1
To adjust the meter to a hanging position follow the steps
shown in Figure 1-3 below.
2. Detach the tilt-stand
maximum reach
1. Extend the tilt-stand until its
4. Re-attach the tilt-stand to an
upright position
this side of the
3. Flip the tilt-stand over until
facing you.
stand is facing the device instead of
Figure 1-3 Tilt-stand at hanging position
Agilent U1251A/U1252A User’s and Service Guide
5
Download from Www.Somanuals.com. All Manuals Search And Download.
1
Getting Started Tutorial
The rotary switch at a glance
6
7
8
5
4
3
2
9
10
1
Figure 1-6 Rotary switch
Table 1-1 Rotary switch description and functions
No.
1
Description/Function
Charge mode (U1252A only) or OFF
2
AC voltage
3
AC voltage, DC voltage or DC+AC voltage (U1252A only)
DC mV, AC mV, AC+DC mV (U1252A only)
Resistance (Ω), Continuity, and Conductance ( nS )
Frequency counter (U1252A only) or Diode
Capacitance or Temperature
4
5
6
7
8
DC µA, AC µA, and AC+DC µA
9
DC mA, DC current, AC mA, AC current, or AC+DC current
10
Square-wave output, Duty cycle, or Pulse width output (for
U1252A) and OFF (for U1251A)
8
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Getting Started Tutorial
1
The operation of each key is shown below. Pressing a key
illuminates a related symbol on the display and emits a beep
from the multimeter. Turning the rotary switch to another
position resets the current operation of the key.
Figure 1-7 below shows the keypad of the U1252A. The ms%
(Pulse width/Duty cycle),
Hz
, and frequency counter
functions are only available on the U1252A.
4
3
5
2
1
6
7
Figure 1-7 U1252A keypad
Table 1-2 Keypad desciption and functions
Button Function when pressed for less than 1 second
Function when pressed for more than 1 second
1
displays the battery capacity for 3 seconds.
Backlit automatically turns off after 30s (default) [1]
.
2
freezes the measured value. In Data Hold
enters Dynamic Recording mode. Press
mode, press again to trigger hold of the next measured again to scroll through Max, Min, Avg, and present
value. In Refresh Hold mode, reading updates
readings (indicated by MAXMINAVG on display).
automatically once reading is stable and count setting
is exceeded [1]
.
3
saves displayed value as a reference to be
subtracted from subsequent measurements. Press
again to see the relative value that has been saved.
enters 1 ms Peak Hold mode. Press
to scroll through Max and Min peak readings.
Agilent U1251A/U1252A User’s and Service Guide
9
Download from Www.Somanuals.com. All Manuals Search And Download.
1
Getting Started Tutorial
Button
Function when pressed for less than 1 second
Function when pressed for more than 1 second
4
scrolls through the measuring function(s) at a
particular rotary switch position.
enters Log Review mode. Press
switch to manual or interval logging data. Press
to
or
to view first or last logged data respectively.
Press
or
to scroll up or down logged data.
Press
for more than 1 second to exit mode.
5
6
scrolls through available measuring ranges
sets to Auto Range mode.
(except when rotary switch is set at TEMP or at Hz [for
U1252A] position) [2]
.
scrolls through available dual-combination
exits Hold, Null, Dynamic Recording, 1 ms
displays (except when rotary switch is set at TEMP, Hz Peak Hold and dual display modes.
or [for U1252A] position, or when meter is in
1 ms peak hold or dynamic recording mode) [3]
.
7
enters Frequency Test mode for current or
enters logging mode. In manual data logging,
to log data manually into memory. In
frequency (Hz), duty cycle (%) and pulse width (ms)
functions. In duty cycle (%) and pulse width (ms)
automatic data logging, data logs automatically [1]
.
Press
for more than 1 second to exit auto data
tests, press
pulse.
to switch to positive or negative logging mode.
Notes for keypad descriptions and functions:
1
See Table 4-1 on page 75, “Available setting options in Setup mode,” for details of available options.
2
3
When rotary switch is at TEMP, press
switch to division of signal frequency by 1 or 100.
When rotary switch is at TEMP, ETC is ON by default. You may press
to switch to ºC or ºF display. When rotary switch is at Hz, press
to
to disable ETC (Environment Temperature
Compensation),
will appear on display. For pulse and duty cycle measurement, press
to switch trigger slope
to positive or negative. When meter is in peak or dynamic-recording mode, press
dynamic recording mode.
to restart 1 ms peak hold or
10
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Getting Started Tutorial
1
The display at a glance
To view the full display (with all segments illuminated),
press and hold the
button while turning the rotary
switch from OFF to any non-OFF position. After you have
finished viewing the full display, press any button to resume
normal functionality based on the rotary switch position.
This is followed by a wake-up feature.
The meter will then enter power save mode once auto power
off (APF) is enabled. To wake the meter up:
1 Turn the rotary switch to OFF position then ON again.
2 Press any button for rotary switch position that is not at
square wave output position. (U1252A only)
3 To set the rotary switch at square wave out position,
press only the Dual, Range and Hold buttons or turn the
rotary switch to another position. (U1252A only)
The LCD signs are explained in the following tables.
Figure 1-8 Display symbols
Agilent U1251A/U1252A User’s and Service Guide
11
Download from Www.Somanuals.com. All Manuals Search And Download.
1
Getting Started Tutorial
Table 1-3 General display symbols
LCD symbol
Description
Remote control
Thermocouple types:
Null math function
(K-type) (J-type)
NULL
Diode / Audible continuity
Audible continuity for resistance
View mode for checking logged data
Data logging indication
Square wave output (U1252A only)
•
•
•
•
Positive slope for pulse width (ms) and duty cycle (%) measurement
Charging capacitor as capacitance measurement
Negative slope for pulse width (ms) and duty cycle (%) measurement
Discharging capacitor as capacitance measurement
Low battery indication
Auto power off enable
Refresh (auto) Hold
Trigger (manual) Hold
TRIG
MAXMINAVG
MAX
Dynamic Recording mode: Present value on primary display
Dynamic Recording mode: Maximum value on primary display
Dynamic Recording mode: Minimum value on primary display
MIN
AVG
Dynamic Recording mode: Average value on primary display
1 ms Peak Hold mode: Positive peak value on primary display
1 ms Peak Hold mode: Negative peak value on primary display
MAX
MIN
12
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Getting Started Tutorial
1
The primary display signs are explained below.
Table 1-4 Priamry display symbols
LCD symbol
Description
Auto range
AC + DC
DC
AC
Polarity, digits, and decimal points for primary display
dBm
dBV
MkHz
MkΩ
nS
Decibel unit relative to 1 mW
Decibel unit relative to 1 V
Frequency units: Hz, kHz, MHz
Resistance units: Ω, kΩ, MΩ
Conductance unit
mV
Voltage units: mV, V
µmA
%
Current units: µA, mA, A
Duty cycle measurement
Pulse width unit
ms
µmnF
ºC
Capacitance units: nF, µF, mF
Celsius temperature unit
ºF
Fahrenheit temperature unit
Percentage scale readout proportional to DC 0–20 mA
Percentage scale readout proportional to DC 4–20 mA
%
%
Agilent U1251A/U1252A User’s and Service Guide
13
Download from Www.Somanuals.com. All Manuals Search And Download.
1
Getting Started Tutorial
The secondary display signs are explained below.
Table 1-5 Secondary display symbols
LCD symbol
Description
AC + DC
DC
AC
Polarity, digits and decimal points for secondary display
Frequency units: Hz, kHz
kHz
No ambient temperature compensation, just thermocouple measurement
Celsius ambient temperature unit
ºC
ºF
Fahrenheit ambient temperature unit
mV
µmA
s
Voltage units: mV, V
Current units: µA, mA, A
Elapsed time unit: s (second) for Dynamic Recording and 1 ms Peak Hold modes
The analog bar emulates the needle on an analog multimeter,
without displaying the overshoot. When measuring peak or
null adjustments and viewing fast-changing inputs, the bar
graph provides a useful indication because it has quicker
update rate for fast response applications.
The bar graph is not used for square wave output, frequency,
duty cycle, pulse width, 4–20 mA% scale, 0–20 mA% scale
and temperature measurements. When frequency, duty cycle,
and pulse width are indicated on the primary display during
voltage or current measurement, the bar graph represents
the voltage or current value. When 4–20 mA% scale or
0–20 mA% scale is indicated on the primary display, the bar
graph represents the current value and not the percentage.
14
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Getting Started Tutorial
1
The “+” or “–” sign is indicated when the positive or
negative value has been measured or calculated. Each
segment represents 2500 or 500 counts depending on the
range indicated on the peak bar graph. See the Table 1-6
below.
Table 1-6 Analog bar range and counts
Range
Counts/segments
2500
Used for the function
V, Ω, Diode
2500
2500
500
V, A, Ω
V, A, Ω, nS
V,
500
500
Agilent U1251A/U1252A User’s and Service Guide
15
Download from Www.Somanuals.com. All Manuals Search And Download.
1
Getting Started Tutorial
Selecting display with the Hz button
The frequency measurement feature helps to detect the
presence of harmonic currents in neutral conductors and
determines whether these neutral currents are the result of
unbalanced phases or non-linear loads. Press
to access
the frequency measurement mode for current or voltage
measurements — voltage or current on the secondary display
and frequency on the primary display. Alternatively, pulse
width (ms) or duty cycle (%) can appear on the primary
display by pressing
again. This enables the
simultaneous monitoring of real-time voltage and current
with frequency, duty cycle, or pulse width. Voltage or
current resumes on primary display after you press and hold
for more than 1 second.
Table 1-7 Selecting display with the Hz button
Rotary switch position (Function)
Primary display
Frequency (Hz)
Pulse width (ms)
Secondary display
AC V
for U1252A
(AC voltage)
Duty cycle (%)
Frequency (Hz)
Pulse width (ms)
Duty cycle (%)
DC V
V for U1251A
for U1252A
(DC voltage)
for U1252A
Frequency (Hz)
Pulse width (ms)
Duty cycle (%)
Frequency (Hz)
Pulse width (ms)
Duty cycle (%)
Frequency (Hz)
Pulse width (ms)
Duty cycle (%)
AC + DC V
AC mV
(AC + DC voltage)
(AC voltage)
(DC voltage)
DC mV
16
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Getting Started Tutorial
1
Table 1-7 Selecting display with the Hz button (continued)
Frequency (Hz)
AC + DC mV
Pulse width (ms)
(AC + DC voltage)
Duty cycle (%)
[for U1252A]
Frequency (Hz)
Pulse width (ms)
Duty cycle (%)
Frequency (Hz)
Pulse width (ms)
Duty cycle (%)
Frequency (Hz)
Pulse width (ms)
Duty cycle (%)
AC µA
DC µA
(AC Current)
(DC current)
AC + DC µA
(AC + DC current)
[for U1252A]
Frequency (Hz)
Pulse width (ms)
Duty cycle (%)
Frequency (Hz)
Pulse width (ms)
Duty cycle (%)
Frequency (Hz)
Pulse width (ms)
Duty cycle (%)
AC mA or A
DC mA or A
(AC current)
(DC current)
AC + DC mA or A
[for U1252A]
(AC + DC current)
Frequency (Hz)
Pulse width (ms)
Duty cycle (%)
- 1 -
Hz (Frequency counter) - press
to select frequency division by 1 [for
U1252A]
Frequency (Hz)
- 100 -
Hz (Frequency counter) - press
to select frequency division by 100 [for
U1252A]
Agilent U1251A/U1252A User’s and Service Guide
17
Download from Www.Somanuals.com. All Manuals Search And Download.
1
Getting Started Tutorial
Selecting display with the Dual button
Press
to select different combinations of dual display.
Normal single display resumes after you press and hold
for more than 1 second. See Table 1-8 below.
Table 1-8 Selecting display with the Dual button
Rotary switch position (Function)
Primary display
Secondary display
Hz (AC coupling)
AC V
AC V
dBm or dBV (select
(AC voltage)
by pressing
)
AC V
AC V
dBm or dBV [1]
Ambient temperature °C or °F
Hz (AC coupling)
AC V
for U1252A
(AC voltage)
AC V
DC V
AC V
Ambient temperature °C or °F
Hz (DC coupling)
DC V
DC V
dBm or dBV [1]
V for U1251A/
for U1252A
DC V
AC V [for U1252A]
Ambient temperature °C or °F
Hz (AC coupling)
AC + DC V
(DC voltage)
DC V
AC + DC V
dBm or dBV [1]
AC + DC V
AC + DC V
AC + DC V
AC mV
for U1252A
(AC + DC voltage)
AC V
DC V
Ambient temperature °C or °F
Hz (AC coupling)
AC mV
dBm or dBV [1]
(AC voltage)
(DC voltage)
AC mV
DC mV
AC mV
Ambient temperature °C or °F
Hz (DC coupling)
DC mV
DC mV
dBm or dBV [1]
DC mV
AC mV
DC mV
Ambient temperature °C or °F
18
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Getting Started Tutorial
1
Table 1-8 Selecting display with the Dual button
AC + DC mV
dBm or dBV
AC + DC mV
AC + DC mV
AC + DC mV
Hz (AC coupling)
AC + DC mV
(AC + DC voltage)
AC mV
DC mV
Ambient temperature °C or °F
Hz (DC coupling)
DC μA
DC μA
DC μA
AC μA
AC μA
AC μA
(DC current)
(AC current)
Ambient temperature °C or °F
Hz (AC coupling)
DC μA
AC μA
Ambient temperature °C or °F
Hz (AC coupling)
AC + DC μA
AC + DC μA
AC + DC μA
AC + DC μA
DC mA
AC μA
DC μA
(AC + DC current)
[for U1252A]
Ambient temperature °C or °F
Hz (DC coupling)
AC mA
DC mA
(DC current)
% (0–20 or 4–20)
DC mA
DC mA
Ambient temperature °C or °F
Hz (AC coupling)
DC mA
AC mA
AC mA
(AC current)
AC mA
Ambient temperature °C or °F
Hz (AC coupling)
AC mA
AC + DC mA
AC + DC mA
AC + DC mA
AC + DC mA
DC A
(AC + DC current)
[for U1252A]
DC mA
Ambient temperature °C or °F
Hz (DC coupling)
(DC current)
(AC current)
DC A
DC A
AC A
AC A
AC A
AC A
Ambient temperature °C or °F
Hz (AC coupling)
DC A
Ambient temperature °C or °F
Agilent U1251A/U1252A User’s and Service Guide
19
Download from Www.Somanuals.com. All Manuals Search And Download.
1
Getting Started Tutorial
Table 1-8 Selecting display with the Dual button
AC + DC A
AC + DC A
Hz (AC coupling)
AC A
(AC + DC current)
[for U1252A]
AC + DC A
DC A
AC + DC A
Ambient temperature °C or °F
Ambient temperature °C or °F
nF / V / Ω / nS
(Capacitance)
(Diode)/
Ω (Resistance)/
nS (Conductance)
TEMP (Temperature)
°C (°F)
°C (°F)
Ambient temperature °C or °F
Ambient temperature °C or °F /
0 °C compensation (select by
pressing
)
Notes for selecting display with the Dual button:
1 Reading of dBm or dBV depends on the last review on AC V. If the last review is in dBV, the following display
will also remain in dBV.
Selection display with the Shift button
The table below shows selection of primary display, with
respect to measuring function (rotary switch position), using
the SHIFT button.
Table 1-9 Selecting display with the Shift button
Rotary switch position (Function)
Primary display
AC V
dBm (in dual display mode) [1]
dBV (in dual display mode) [1]
DC V
(AC Voltage)
V for U1251A
DC V
AC V
for U1252A
(AC + DC Voltage)
AC + DC V
20
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Getting Started Tutorial
1
Table 1-9 Selecting display with the Shift button
DC mV
AC mV
for U1252A
(AC + DC Voltage)
AC + DC mV
Ω
Ω
nS
Ω
(Resistance)
Diode
Hz
(Diode Test and Frequency)
/ TEMP
Capacitance
Temperature
(Capacitance and Temperature)
DC μA
AC μA
AC + DC μA [for U1252A]
(AC Current)
DC mA
AC mA
AC + DC mA
%(0–20 or 4–20)
DC A
(DC Current)
AC A
AC + DC A [for U1252A]
(AC + DC Current)
Duty cycle (%)
Pulse width (ms)
(Square wave output for U1252A)
Notes for selecting display with the Shift button:
1 Press
to switch between dBm and dBV measurement.
Press
for more than 1 second to return to AC V measurement only.
Agilent U1251A/U1252A User’s and Service Guide
21
Download from Www.Somanuals.com. All Manuals Search And Download.
1
Getting Started Tutorial
The terminals at a glance
To avoid damaging the multimeter, do not exceed the input limit.
WARNING
Figure 1-9 U1252A connector terminals
Table 1-10 Terminal connections for different measuring functions
Rotary switch position
Input terminals
V . mV . Ω .
Overload protection
COM
1000 V R.M.S.
for U1252A
V for U1251A
.TEMP
1000 V R.M.S. for short circuit <0.3 A
Ω
μA . mA
COM
COM
440 mA / 1000 V 30 kA fast-acting
fuse
A
11 A / 1000 V 30 kA fast-acting fuse
440 mA / 1000 V fast-acting fuse
for U1252A
COM
COM
22
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent U1251A and U1252A Handheld Digital Multimeter
User’s and Service Guide
2
Measuring Temperature 43
Alerts and Warning During Measurement 47
Overload alert 47
Input warning 47
Charge terminal alert 48
This chapter contains information on how to make
measurements using the U1251A and U1252A handheld
digital multimeter.
Agilent Technologies
23
Download from Www.Somanuals.com. All Manuals Search And Download.
2
Making Measurements
Understanding The Measurement Instructions
When making measurements, follow the numerical steps
labelled in the diagrams. Refer to Table 2-1 below for a
description of the steps.
Table 2-1 Numerical steps descriptions
No.
Instructions
Turn the rotary switch to the measurement option shown in the diagram
Connect the test leads into the input terminals shown in the diagram
Probe the test points
Read the results on the display
Measuring Voltage
The U1251A and U1252A offer true-RMS readings for AC
measurements that are accurate for sine waves, square
waves, triangle waves, staircase waves, and other waveforms
without any DC offset.
For AC with DC offset, use AC + DC measurement on
or
rotary switch location. This applies only to
U1252A.
Ensure that the terminal connections are correct for that particular
measurement before proceeding with the measurement. To avoid
damaging the device, do not exceed the input limit.
WARNING
24
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Making Measurements
2
Measuring AC voltage
Set up the multimeter to measure AC voltage as shown in
Figure 2-1. Probe the test points and read the display.
Red Lead
Figure 2-1 Measuring AC voltage
Press
to display frequency on secondary display. See Table 1-8 of
NOTE
“Selecting display with the Dual button” on page 18 for a list of the
different combinations available on the secondary display.
Agilent U1251A/U1252A User’s and Service Guide
25
Download from Www.Somanuals.com. All Manuals Search And Download.
2
Making Measurements
Measuring DC voltage
Set up the multimeter to measure DC voltage as shown in
Figure 2-2. Probe the test points and read the display.
MAX
Red Lead
Black Lead
Figure 2-2 Measuring DC voltage
26
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Making Measurements
2
Measuring Current
µA & mA measurement
Set up the multimeter to measure μA & mA as shown in
Figure 2-3. Probe the test points and read the display.
•
•
Press
if necessary to ensure
is shown on the display.
NOTE
For μA measurement, set the rotary switch to
the positive test lead to μA.mA.
and connect
•
•
•
connect the positive test lead to μA.mA.
and
For A (ampere) measurement, set the rotary switch to
connect the positive test lead to A.
and
Press
to display dual measurements. See Table 1-8 of “Selecting
display with the Dual button” on page 18 for a list of dual
measurements available. Press and hold
to exit dual display mode.
for more than 1 second
Agilent U1251A/U1252A User’s and Service Guide
27
Download from Www.Somanuals.com. All Manuals Search And Download.
Making Measurements
2
Percentage scale of 4 mA to 20 mA
Set up the multimeter to measure percentage scale as shown
in Figure 2-4. Probe the test points and read the display.
•
•
Press
to select percentage scale display. Ensure that
or
NOTE
is shown on the display.
The percentage scale for 4 mA to 20 mA or 0 mA to 20 mA is calculated
using its corresponding DC mA measurement. The U1251A/U1252A
will automatically optimize the best resolution according to Table 2-2
below.
•
Press
to change the measurement range.
The percentage scale for 4 mA to 20 mA or 0 mA to 20 mA
is set to two ranges as follows:
Table 2-2 Percentage scale and measurement range
Percentage scale (4 mA to 20 or 0 mA to 20 mA)
Always auto range
DC mA auto or manual
range
999.99%
9999.9%
50 mA, 500 mA
Agilent U1251A/U1252A User’s and Service Guide
29
Download from Www.Somanuals.com. All Manuals Search And Download.
Making Measurements
2
A (ampere) measurement
Set up the multimeter to measure A (ampere) as shown in
Figure 2-5. Probe the test points and read the display.
Connect the red and black test leads to the 10 A input terminal A and COM
respectively. The meter is set to A measurement automatically when the
red test lead is plugged into the A terminal.
NOTE
Black Lead
Red Lead
Figure 2-5 A (ampere) current measurement
Agilent U1251A/U1252A User’s and Service Guide
31
Download from Www.Somanuals.com. All Manuals Search And Download.
2
Making Measurements
Frequency Counter
•
•
Use the frequency counter for low voltage applications only. Never use
the frequency counter for line power system.
WARNING
measurement mode available under the current or voltage
measurement instead of frequency counter.
Set up the multimeter to measure frequency as shown in
Figure 2-6. Probe the test points and read the display.
•
Press
to select the Frequency counter (Hz) function. “-1-“ on
NOTE
the secondary display means the input signal frequency is divided by 1.
This allows signals of up to a maximum frequency of 985 kHz to be
measured.
•
If the reading is unstable or is zero, press
to select the division
of the input signal frequency by 100. This allows for a higher frequency
range of up to 20 MHz to be measured.
•
•
The signal is out of range if the reading is still unstable after the above
step.
While the secondary display shows “-1-“, you may scroll through the
pulse width (ms), duty cycle (%), and frequency (Hz) measurements by
pressing
.
32
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
2
Making Measurements
Measuring Resistance, Conductance and Testing Continuity
CAUTION
before measuring the resistance to prevent any possible damage to the
multimeter or the device under test.
Set up the multimeter to measure resistance as shown in
Figure 2-7. Then probe the test points (by shunting the
resistor) and read the display.
nS
Ω
Black Lead
Red Lead
Figure 2-7 Measuring resistance
34
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Making Measurements
2
Press
to scroll through audible continuity, conductance
and resistance tests as shown in Table 2-8.
Press
Audible
continuity
Press
Press
Figure 2-8 Audible continuity, conductance, and resistance test.
Agilent U1251A/U1252A User’s and Service Guide
35
Download from Www.Somanuals.com. All Manuals Search And Download.
2
Making Measurements
Audible continuity
In the range of 0–500 Ω, the beeper will sound if the
resistance value falls below 10 Ω. For other ranges, the
beeper will sound if the resistance falls below the typical
values indicated in Table 2-3 below.
Table 2-3 Audible continuity measurement range
Measurement range
500.00 Ω
Beeper sound threshold
< 10 Ω
5.0000 kΩ
< 100 Ω
50.000 kΩ
< 1 kΩ
500.00 kΩ
< 10 kΩ
5.0000 MΩ
50.000 MΩ
< 100 kΩ
< 1 MΩ
< 10 MΩ
Conductance
Set up the multimeter to measure conductance as shown in
Figure 2-9. Probe the test points and read the display.
very high resistance of up to 100 GΩ.
As the high-resistance readings are susceptible to noise, you
can capture the average readings by using the Dynamic
Recording mode. Refer to the section “Dynamic
Recording” on page 50 for more information.
36
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
2
Making Measurements
Testing Diodes
Disconnect the circuit power and discharge all the high-voltage capacitors
CAUTION
To test a diode, turn the power off to the circuit and remove
the diode from the circuit. Set up the multimeter as shown
in Figure 2-10, then use the red probe lead on the positive
terminal (anode) and use the black probe lead on the
negative terminal (cathode) and read the display.
•
•
The cathode is the side with band(s).
NOTE
The meter can display the diode’s forward bias of up to approximately
2.1 V. A typical diode’s forward bias is between the range of 0.3 V to
0.8 V.
Next, reverse the probes and measure the voltage across the
diodes again as shown in Figure 2-11 on page 40. The
diode’s test result is based on the following:
• The diode is considered good if the meter displays “OL”
in reverse bias mode.
• The diode is considered shorted if the meter displays
approximately 0 V in both forward and reverse bias
modes, and the meter beeps continuously.
• The diode is considered open if the meter displays “OL”
in both forward and reverse bias modes.
38
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Making Measurements
2
Measuring Capacitance
Disconnect the circuit power and discharge all the high-voltage capacitors
before measuring the capacitance to prevent any possible damage to the
meter or the device under test. To confirm that the capacitors have
discharged, use the DC voltage function.
CAUTION
The meter measures capacitance by charging the capacitor
with a known current for a period of time, measuring the
voltage and then calculating the capacitance. The larger the
capacitor, the longer the charge time. Below are some tips
for measuring capacitance:
• For measuring capacitance greater than 10,000 µF,
discharge the capacitor first, then select a suitable
range for the measurement. This will speed up
measuring time in order to obtain the correct
capacitance value.
• For measuring smaller capacitance ranges, press
with the test leads open to subtract any residual
capacitance from the meter or the leads.
means the capacitor is charging.
discharging.
means the capacitor is
NOTE
Set up the multimeter as shown in Figure 2-12. Use the red
probe lead on the positive terminal of the capacitor and the
black probe lead on the negative terminal and read the
display.
Agilent U1251A/U1252A User’s and Service Guide
41
Download from Www.Somanuals.com. All Manuals Search And Download.
Making Measurements
2
Measuring Temperature
Do not bend the thermocouple leads at sharp angles. Repeated bending
over a period of time may break the leads.
CAUTION
The bead type thermocouple probe is suitable for making
temperature measurements between –20 °C to 200 °C in
PTFE compatible environments.
Do not use the bead-type thermocouple probe beyond the
recommended operating temperature range. Do not immerse
thermocouple probe designed for each application — an
immersion probe for liquid or gel and an air probe for air
measurements.
Figure 2-15 or observe the following steps:
1 Press
to select the temperature measurement.
2 Connect the miniture thermal probe into the
non-compensation transfer adaptor as shown in
Figure 2-13.
3 Connect the thermal probe with the adapter into the
meter input terminals as shown in Figure 2-14.
4 For best performance, place the meter in the operating
environment for at least one hour to stabilize the unit to
environment temperatures.
5 Clean the measurement surface and make sure the probe
is securely touching the surface. Remember to disable the
applied power.
6 When measuring above the ambient temperature, move the
thermocouple along the surface until you get the highest
temperature reading.
7 When measuring below ambient temperature, move the
thermocouple along the surface until you get the lowest
temperature reading.
Agilent U1251A/U1252A User’s and Service Guide
43
Download from Www.Somanuals.com. All Manuals Search And Download.
2
Making Measurements
8 For a quick measurement, use the 0 °C compensation
adapter to see the temperature variation of the
thermocouple sensor. The 0 °C compensation adapter
assists in measuring the relative temperature immediately.
Figure 2-13 Connecting the thermal probe into the non-compensation
transfer adapter
Figure 2-14 Connecting the probe with adapter into the multimeter
44
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Making Measurements
2
If you are working in a constantly varying environment,
where ambient temperatures are not constant, do the
following:
1 Press
to select 0 °C compensation. This gives a
quick measurement of the relative temperature.
2 Avoid contact between the thermocouple probe and the
measurement surface.
3 After a constant reading is obtained, press
to set
the reading as the relative reference temperature.
4 Touch the measurement surface with the thermocouple
probe.
5 Read the display for the relative temperature.
Agilent U1251A/U1252A User’s and Service Guide
45
Download from Www.Somanuals.com. All Manuals Search And Download.
Making Measurements
2
Alerts and Warning During Measurement
Overload alert
For your safety, look out for this alert. When you are alerted,
immediately remove the test leads from the measuring source.
WARNING
The meter provides an overload alert for voltage
measurements in both auto and manual range modes. The
meter beeps periodically once the measuring voltage exceeds
1010 V. For your safety, please be aware of this alert.
Input warning
The meter sounds an alert beep when the test lead is
inserted into the A input terminal but the rotary switch is
not set to the corresponding mA.A location. The primary
display indicates a flashing “A- Err” until the test lead is
removed from the A input terminal. Refer to Figure 2-16.
Figure 2-16 Input terminal warning
Agilent U1251A/U1252A User’s and Service Guide
47
Download from Www.Somanuals.com. All Manuals Search And Download.
2
Making Measurements
Charge terminal alert
terminal
detects a voltage level of more than 5 V and the rotary
switch is not set to the corresponding
location. The
primary display indicates a flashing “Ch.Err” until the lead
is removed from the
input terminal.
Refer to Figure 2-17 below.
Figure 2-17 Charge terminal alert
48
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent U1251A and U1252A Handheld Digital Multimeter
User’s and Service Guide
3
Manual logging 61
Interval logging 63
Reviewing logged data 65
Square Wave Output (for U1252A) 67
Remote Communication 71
This chapter contains information on the functions and
features that are available for the U1251A and U1252A
digital multimeter.
Agilent Technologies
49
Download from Www.Somanuals.com. All Manuals Search And Download.
3
Features and Functions
Dynamic Recording
The Dynamic Recording mode can be used to detect
intermittent turn-on or turn-off voltage, current surges or to
verify measurement performance without you being present
during the process. While the readings are being recorded,
you are free to perform other tasks.
The average reading is useful for smoothing out unstable
inputs, estimating the percentage of time a circuit is
operating and verifying circuit performance. The elapsed
time is shown on the secondary display. The maximum time
is 99999 seconds. When this maximum time is exceeded, “OL”
is shown on the display.
1 Press
for more than 1 second to enter Dynamic
Recording mode. The meter is now in continuous mode or
non-data hold (non-trigger) mode. “MAXMINAVG” and
the present value of measurement are displayed. The
multimeter will beep when a new maximum or minimum
value is recorded.
2 Press
to cycle through maximum, minimum,
average, and present readings. The MAX, MIN, AVG, and
MAXMINAVG indicators will appear, corresponding to the
displayed readings.
3 Press
or
for more than 1 second to exit
Dynamic Recording mode.
•
•
Press
to restart the dynamic recording.
NOTE
The average value is the true average of all the measured values taken
in the Dynamic Recording mode. If an overload is recorded, the
averaging function will stop and the average value becomes
"OL"(overload).
is disabled in Dynamic Recording mode.
50
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Features and Functions
3
Press Max Min
for > 1 sec
Press Max Min
Press Max Min for > 1 sec
or press
for > 1 sec
Press Max Min
Press Max Min for > 1 sec
or press for > 1 sec
Press Max Min
Press Max Min for > 1 sec
or press
for > 1 sec
Figure 3-1 Dynamic recording mode operation
Agilent U1251A/U1252A User’s and Service Guide
51
Download from Www.Somanuals.com. All Manuals Search And Download.
3
Features and Functions
Data Hold (Trigger Hold)
The data hold function allows the user to freeze the
displayed digital value.
1 Press to freeze the displayed value and to enter the
manual trigger mode. TRIG
is displayed.
2 Press
again to freeze the next value being
measured. TRIG flashes before the new value is updated
on the display.
3 Press and hold
or
for more than 1 second to
quit the data hold function.
Press
Press
Press
Figure 3-2 Data hold mode operation
52
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Features and Functions
3
Refresh Hold
The Refresh Hold function allows you to freeze the displayed
value. The bar-graph is not held and will continue to reflect
the instantaneous measured value. You can use the Setup
mode to enable Refresh Hold mode when you are working
with fluctuating values. This function will auto trigger or
update the held value with a new measured value and emit
a tone as a reminder.
1 Press
to enter Refresh Hold mode. The present
symbol will appear.
value will be held and the
2 It will be ready to hold a new measured value once the
variation of measured values exceeds the variation count
setting. While the multimeter is waiting for a new stable
value, the
symbol will flash.
3 The
symbol will stop flashing once the new
measured value is stable and the new value will be
updated to the display. The symbol will remain on and
the multimeter will emit a tone to remind you of this.
4 Press
again to quit the Refresh Hold function.
Agilent U1251A/U1252A User’s and Service Guide
53
Download from Www.Somanuals.com. All Manuals Search And Download.
3
Features and Functions
Press
Variation counts
> setting
The HOLD sign flashes
until the measured
value is stable and
updated
Measurement
is updated
Press
to return
Figure 3-3 Refresh hold mode operation
•
•
•
For voltage and current measurements, the holding value will not be
updated if the reading is below 500 counts.
NOTE
For resistance and diode measurements, the holding value will not be
updated if the reading is in “OL” (open state).
The holding value may not be updated when the reading does not reach
a stable state for all measurements.
54
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Features and Functions
3
Null (Relative)
The Null function subtracts a stored value from the present
measurement and displays the difference between the two.
1 Press
to store the displayed reading as the
reference value to be subtracted from subsequent
measurements and to set the display to zero. NULL is
displayed.
2 Press
to see the stored reference value. NULL
flashes for 3 seconds before the display returns to zero.
3 To exit this mode, press
while NULL is flashing on
the display.
•
•
Null can be set for both auto and manual range settings, but not in the
event of an overload.
NOTE
When taking a resistance measurement and the meter reads a
non-zero value due to the presence of test leads, use the Null function
to adjust the display to zero.
•
When taking a DC voltage measurement, the thermal effect will
influence the accuracy. Short the test leads and press Null once the
displayed value is stable in order to zero out the display.
Agilent U1251A/U1252A User’s and Service Guide
55
Download from Www.Somanuals.com. All Manuals Search And Download.
Features and Functions
3
Decibel Display
The dBm operation calculates the power delivered to a
reference resistance relative to 1 mW and can be applied to
DC V, AC V, and AC + DC V measurements for decibel
conversion. The voltage measurement is converted to dBm by
using the following formula:
The reference resistance may be selected from 1~9999 Ω in
Setup mode. The default value is 50 Ω.
The decibel of voltage is calculated with respect to 1 V. The
equation for the voltage measurement is as follows:
dBV = 20 log Vin
10
1 At the
,
, or
rotary switch position,
press
to scroll to the dBm measurement on the
primary display. The secondary display indicates the AC
voltage measurement.
If the rotary switch is at the “~ V” position, press
to switch
NOTE
between the dBV and the dBm measurements. The dBm or the dBV
measurements can be selected at the ACV position. The selection will be
the reference for other voltage measurements.
2 Press
for more than 1 second to exit this mode.
Agilent U1251A/U1252A User’s and Service Guide
57
Download from Www.Somanuals.com. All Manuals Search And Download.
Features and Functions
3
1 ms Peak Hold
The Peak Hold function allows the measurement of peak
voltage for analysis of components such as power
distribution transformers and power factor correction
capacitors. The peak voltage obtained can be used to deter-
mine the crest factor:
Crest factor = Peak value/True RMS value
1 Press
for more than 1 second to toggle 1 ms Peak
Hold mode ON / OFF.
2 Press
to scroll through maximum and minimum
MAX indicates the maximum peak,
MIN indicates the minimum peak.
peak readings.
while
•
•
If the reading is "OL", press
to re-start the peak-recording measurement.
to change the measuring range and
NOTE
If you need to re-start the peak recording, press
.
3 Press and hold
or
for more than 1 second
to exit this mode.
4 According to the measurements shown in Figure 3-6 on
page 60, the crest Factor will be 2.5048/1.768 =1.416.
Agilent U1251A/U1252A User’s and Service Guide
59
Download from Www.Somanuals.com. All Manuals Search And Download.
Features and Functions
3
Data Logging
The data logging function provides the convenience of
recording test data for future review or analysis. Since data
is stored in nonvolatile memory, the data remains saved
even if the multimeter is turned OFF or the battery is
changed.
The two options offered are manual (hand) logging and
interval (time) logging functions, which is determined in the
Setup mode.
Data logging records the values on the primary display only.
To use the data logging function, you will need to connect the multimeter
to a PC using the U1173A IR-to-USB cable (purchased separately) and
download the data logging software from Agilent’s website. Please go to:
http://www.agilent.com/find/hhTechLib to download the software.
NOTE
Manual logging
Firstly, ensure that manual (hand) logging is specified in
Setup mode.
1 Press
for more than 1 second to store the present
value and function on the primary display to the
non-volatile memory. The
and the logging index
will be indicated. The logging index flashes on the
secondary display for 3 seconds before returning to
normal display.
2 Press and hold
again for the next value that you
would like to save into the memory.
Agilent U1251A/U1252A User’s and Service Guide
61
Download from Www.Somanuals.com. All Manuals Search And Download.
3
Features and Functions
Press
Press
Figure 3-7 Hand (Manual) logging mode operation
The maximum data that can be stored is 100 entries. When the 100 entries
are filled, the secondary display indicates “FULL”, as shown in Figure 3-8.
NOTE
Press
Figure 3-8 Full Log
62
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Features and Functions
3
Interval logging
Firstly, ensure that interval (time) logging is specified in
Setup mode.
1 Press
value and function on the primary display into the
non-volatile memory. The and the logging index will
for more than 1 second to store the present
be indicated. The readings are automatically saved into
the permanent memory at intervals set using the Setup
mode.
The maximum data that can be stored is 200 entries. When the 200 entries
are filled, “FULL” is indicated on the secondary display.
NOTE
NOTE
2 Press
for more than 1 second to exit this mode.
When interval (automatic) logging is enabled, all keypad operations are
disabled, except for the Log function.
Agilent U1251A/U1252A User’s and Service Guide
63
Download from Www.Somanuals.com. All Manuals Search And Download.
Features and Functions
3
Reviewing logged data
1 Press
for more than 1 second to enter the Log
Review mode. The last recorded entry and the last logging
index are displayed.
2 Press
(automatic) logging review mode.
3 Press to ascend or to descend through the logged
data. Press to select the first record and press to
select the last record for quick navigation.
to switch between hand (manual) and interval
4 Press
for more than 1 second at the respective Log
for more than 1 second to exit mode.
Review mode to clear logged data.
5 Press
6 During the data review in either manual or interval
logging mode, press
for more than one second to
clear all the logged values.
Agilent U1251A/U1252A User’s and Service Guide
65
Download from Www.Somanuals.com. All Manuals Search And Download.
Features and Functions
3
Square Wave Output (for U1252A)
The square wave output function can be used to generate a
PWM (pulse width modulation) output or provide a synchro-
nous clock source (baud rate generator). You can also use
this function to check and calibrate flow-meter displays,
counters, tachometers, oscilloscopes, frequency converters,
frequency transmitters, and other frequency input devices.
1 Turn the rotary switch to
position. The default
display setting is 600 Hz on the secondary display and
50% duty cycle on the primary display.
2 Press
or
to scroll through the available frequencies
(choose from the available 28 frequencies):
Frequency (Hz)
0.5, 1, 2, 5, 10, 15, 20, 25, 30, 40, 50, 60, 75, 80, 100, 120, 150, 200, 240, 300, 400, 480, 600, 800, 1200, 1600,
2400, 4800
Pressing
is the same as pressing
.
NOTE
3 Press
to select the duty cycle (%) on the primary
to adjust the duty cycle. Duty cycle can
display.
4 Press
or
be set for 256 steps and each step is 0.390625%. The
display only indicates the best resolution with 0.001%.
Agilent U1251A/U1252A User’s and Service Guide
67
Download from Www.Somanuals.com. All Manuals Search And Download.
3
Features and Functions
Figure 3-11 Frequency adjustment for square wave output
5 Press
display.
6 Press
to select the pulse width (ms) on the primary
or to adjust the pulse width. The pulse
width can be set for 256 steps and each step is 1/ (256 x
Frequency). The display range automatically adjusts in the
range of 9.9999~9999.9 ms.
68
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Features and Functions
3
Remote Communication
The meter has a bi-directional (full duplex) communication
capability that speeds data storing from the meter to the PC.
To use this feature, you are required to use the optional
IR-USB cable with the application software included in the
CD or downloadable from the Agilent Web site.
For details on performing PC to meter remote
communication click on Help after launching the Agilent GUI
Data Logger Software or refer to the GUI Data Logger Quick
Connect
to PC
Figure 3-14 Cable connection for remote communication
Agilent U1251A/U1252A User’s and Service Guide
71
Download from Www.Somanuals.com. All Manuals Search And Download.
3
Features and Functions
72
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent U1251A and U1252A Handheld Digital Multimeter
User’s and Service Guide
4
Setting Parity Check 90
Setting Data Bit 91
Setting Echo Mode 92
Setting Print Mode 93
Returning to Default Factory Settings 94
This chapter shows you how to change the default factory
settings of the U1251A and U1252A and other available
setting options
Agilent Technologies
73
Download from Www.Somanuals.com. All Manuals Search And Download.
4
Changing The Default Setting
Selecting Setup Mode
To enter the Setup mode, perform the following steps:
1 Turn the meter OFF.
2 From the OFF position, press and hold
while
turning the rotary switch to any non-OFF position.
When you hear a beep, the meter is in Setup mode and you can release
NOTE
To change a menu item setting in Setup mode, perform the
following steps:
1 Press
2 Press
or
or
to scroll through the menu items.
to scroll through the available settings.
See Table 4-1, “Available setting options in Setup mode,”
for details of available options.
3 Press
to save changes. These parameters will remain
in the non-volatile memory.
4 Press
for more than 1 second to exit Setup mode.
74
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Changing The Default Setting
4
Table 4-1 Available setting options in Setup mode
Menu Item
Available setting options
Description
Default factory
setting
Display
Description
Display
OFF
Enables Data Hold (manual trigger)
rHoLd [1]
500
Refresh Hold
Sets the variation count that determines Refresh
Hold (auto trigger)
100–1000
Enables the reset of factory settings by pressing
rESEt
dEFAU
dEFAU
Reset
and holding
for more than 1 second
Enables auto send of data to the PC continuously
when set to ON
Print
ECHO
dAtAb
PArtY
ON, OFF
ON, OFF
OFF
OFF
Print
Echo
Enables the return of characters to the PC when
set to ON
Sets the data bit length for the remote
communication (remote control with PC)
7-bit, 8-bit
En, Odd, nOnE
8-bit
nOnE
Data bits
Parity check
Sets the even, odd or no parity check for the
remote communication (remote control with PC)
2400 Hz, 4800
Hz, 9600 Hz,
19200 Hz
Sets the baud rate for the remote communication
(remote control with PC)
bAUd
b-Lit
9600 Hz
30 s
Baud rate
Sets the timer for the auto turn-off for the backlit
display
[2]
1–99 s
Backlit display
OFF
Disables the auto turn-off for the backlit display
2400 Hz, 1200
Hz, 600 Hz,
300 Hz
Sets the frequency of the meter’s beep sound
Frequency of beep
sound of meter
bEEP
2400 Hz
OFF
Disables the meter’s beep sound
Sets the % scale readout
0–20mA, 4–20
mA
PErnt
APF
4–20 mA
10 m
Percentage scale
Auto power off
[2]
Sets the timer for auto power off
Disables auto power off
1–99 m
OFF
0.5 Hz, 1 Hz, 2
Hz, 5 Hz
Minimum frequency
that can be measured
Sets the minimum frequency that can be
measured
FrEq
rEF
0.5 Hz
50 Ω
Reference impedance
for dBm measurement
Sets the reference impedance for dBm
measurement
[2]
1–9999 Ω
Agilent U1251A/U1252A User’s and Service Guide
75
Download from Www.Somanuals.com. All Manuals Search And Download.
4
Table 4-1 Available setting options in Setup mode (continued)
tYPE K
tYPE J
Hand
Sets the thermocouple type to K-type
[3]
tYPE K
Hand
t.CoUP
Thermocouple
Data logging
Sets the thermocouple type to J-type
Enables manual data logging
d-LoG
[2]
Sets the interval for automatic data logging
Sets the temperature measurement to °C but
1–9999 s
d-CF
pressing
swaps the display to °F
d-F
Sets the temperature measurement to °F
[4]
d-CF
tEMP
Temperature
Sets the temperature measurement to °F but
pressing
d-FC
d-C
swaps display to °C
Sets the temperature measurement to °C
Notes for setting options in Setup mode:
1
2
3
4
This is the first option displayed once the user enters Setup mode.
For b-Lit, APF, rEF and d-LoG menu items, the user can select the digit to be adjusted by pressing
This menu option is only available for the U1252A.
To view the tEMP menu item, press
for more than 1 second.
76
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Changing The Default Setting
4
Setting Data Hold/Refresh Hold Mode
1. Set OFF to enable the Data Hold mode (manual trigger by
key or bus via the remote control).
2. Set the variation count within the 100 to 1000 range to
enable the Refresh Hold mode (auto trigger). When the
variation of the measuring value exceeds the setting of the
variation count the Refresh Hold will be ready to trigger.
Press
Press
Press
Press
Press
Figure 4-1 Data hold/Refersh hold setup
Agilent U1251A/U1252A User’s and Service Guide
77
Download from Www.Somanuals.com. All Manuals Search And Download.
4
Changing The Default Setting
Setting Data Logging Mode
1 Set to “Hand” to enable the manual data logging mode.
2 Set the interval within 0001 to 9999 seconds to enable
interval (automatic) data logging mode.
3 Press and hold
or
for more than 1 second to switch
between manual and interval data logging setup.
Press
Press
Press
Press
Press
Press
Figure 4-2 Data logging setup
78
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Changing The Default Setting
4
Setting Thermocouple Types (U1252A only)
The thermocouple sensor types that can be selected are
type-K (default) or type-J. Press
between the J and the K type.
or
to switch
Press
Figure 4-3 Thermocouple type setup
Agilent U1251A/U1252A User’s and Service Guide
79
Download from Www.Somanuals.com. All Manuals Search And Download.
4
Changing The Default Setting
Setting Reference Impedance for dBm Measurement
The reference impedance can be set from 1 to 9999 W. The
default value is 50 W.
Press
Press
Press
Press
Figure 4-4 Reference impedance for dBm measurement setup
80
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Changing The Default Setting
4
Setting Minimum Frequency Measurement
The minimum frequency setup influences the measuring
rates for frequency, duty cycle, and pulse width. The typical
measuring rate defined at the general specification is based
on the minimum frequency of 1 Hz.
Press
Press
Press
Press
Press
Figure 4-5 Minimum frequency setup
Agilent U1251A/U1252A User’s and Service Guide
81
Download from Www.Somanuals.com. All Manuals Search And Download.
4
Changing The Default Setting
Setting Temperature Unit
Four combination displays are available:
• Celsius only (°C on primary display) single display setting
• Celsius-Fahrenheit (d-CF) and Fahrenheit-Celsius (d-FC)
dual display setting.
Primary-Secondary Display can be swapped by pressing
.
NOTE
• Fahrenheit only (°F on primary display) single display
setting.
82
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
4
Changing The Default Setting
Setting Auto Power Saving Mode
• The timer for APF (Auto Power OFF) can be set for the
range of 1 to 99 minutes.
• To activate the meter after it has “auto powered-off”, turn
the rotary switch to the OFF position. Then turn it back
on again.
•
will be shown on the display during subsequent
measurements.
84
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
4
Changing The Default Setting
Setting Percentage (%) Scale Readout
This setting converts the DC current measuring display to
percentage (%) scale readout — 4—20 mA or 0—20 mA as
proportional to 0~100%. The 25% scale readout represents
DC 8 mA at 4—20 mA and DC 5 mA at 0—20 mA.
Press
Figure 4-8
% scale readout setup
86
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Changing The Default Setting
4
Setting Beep Frequency
The driving frequency can be set to 2400, 1200, 600,
or 300Hz. “OFF” disables the beep.
Press
Press
Press
Press
Press
Press
Figure 4-9 Beep frequency setup
Agilent U1251A/U1252A User’s and Service Guide
87
Download from Www.Somanuals.com. All Manuals Search And Download.
4
Changing The Default Setting
Setting Backlight Timer
• The timer can be set to 1 to 99 seconds. The backlight
will turn off automatically after the set period.
• “0FF” disables turning off backlit automatically.
Press
Press
Press
Press
Press
Figure 4-10 Backlit timer setup
88
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Changing The Default Setting
4
Setting Baud Rate
Select the baud rate for remote control. The available
settings are 2400, 4800, 9600, and 19200 Hz.
Press
Press
Press
Press
Press
Figure 4-11 Baud rate setup remote control
Agilent U1251A/U1252A User’s and Service Guide
89
Download from Www.Somanuals.com. All Manuals Search And Download.
4
Changing The Default Setting
Setting Parity Check
Select the parity check for remote control. It can be set to
none, even, or odd bit.
Press
Press
Press
Press
Figure 4-12 Parity check setup
90
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
4
Changing The Default Setting
Setting Echo Mode
• Echo ON enables the return of characters to the PC in
remote communication.
• Echo OFF disables the return of characters to the PC.
Press
Figure 4-14 Echo mode setup for remote control
92
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Changing The Default Setting
4
Setting Print Mode
Print ON enables the printing of measured data to the PC
when the measuring cycle is completed. In this mode, the
meter automatically sends the newest data to the host
continuously but does not accept any commands from the
host.
flashes during the Print operation.
Press
Figure 4-15 Print mode setup for remote control
Agilent U1251A/U1252A User’s and Service Guide
93
Download from Www.Somanuals.com. All Manuals Search And Download.
4
Changing The Default Setting
Returning to Default Factory Settings
• Press
for more than 1 second to reset to the
default factory settings all menu options except the
Temperature setting.
• The Reset menu item automatically reverts to Refresh
Hold menu item after reset has taken place.
Press
for > 1 sec
Figure 4-16 Reset setup
94
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent U1251A and U1252A Handheld Digital Multimeter
User’s and Service Guide
5
Fuse checking procedure 105
Replacing the fuse 106
Troubleshooting 108
Replaceable Parts 109
To order replaceable parts 109
This chapter will go through how to troubleshoot the
handheld digital multimeter if any problems arise.
Agilent Technologies
95
Download from Www.Somanuals.com. All Manuals Search And Download.
5
Maintenance
Introduction
Any repair or service which is not covered in this manual should only
be performed by qualified personnel.
CAUTION
General maintenance
Ensure that the terminal connections are correct for that particular
measurement before proceeding. To avoid damaging the device, do not
exceed the input limit.
WARNING
Besides the above hazard, dirt, or moisture in the terminals
can distort readings. The steps for cleaning are as follows:
1 Turn the meter off and remove the test leads.
2 Turn the meter over and shake out any dirt that may
have accumulated in the terminals.
3 Wipe the case with a damp cloth and a mild detergent —
do not use abrasives or solvents. Wipe the contacts in
each terminal with a clean swab dipped in alcohol.
Battery replacement
The meter is powered by a 9 V Ni-MH rechargeable battery,
Figure 5-1 below).
To ensure that the correct battery type is used, replace the
battery immediately when the low battery sign flashes. If
your meter has the rechargeable battery type, refer to the
section “Charging the battery” on page 98. The steps for
battery replacement are listed below:
96
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Maintenance
5
Figure 5-1 9 V rectangular battery
1 At the rear panel, turn the screw on the battery cover
from the LOCK to the OPEN position (counterclockwise).
2 Slide down the battery cover.
3 Lift the battery cover up.
4 Replace the specified battery.
5 Reverse the above steps to close the cover.
List of compatible batteries for the Agilent U1251A :
NOTE
• 9 V Alkaline non-chargeable battery (ANSI/NEDA 1604A or IEC 6LR61)
• 9 V Carbon-zinc non-chargeable battery (ANSI/NEDA 1604D or IEC6F22)
List of of compatible batteries for the Agilent U1252A:
• 9 V size 300 mAH Ni-MH rechargeable battery, 7.2 V nominal voltage, or
8.4 V nominal voltage
• 9 V Alkaline non-chargeable battery (ANSI/NEDA 1604A or IEC 6LR61)
• 9 V Carbon-zinc non-chargeable battery (ANSI/NEDA 1604D or IEC6F22)
Agilent U1251A/U1252A User’s and Service Guide
97
Download from Www.Somanuals.com. All Manuals Search And Download.
5
Maintenance
Storage considerations
To avoid instrument damage from battery leakage:
CAUTION
•
•
Always remove dead batteries immediately.
It is recommended that the battery is removed and stored separately if
the multimeter is to be unused for long periods of time.
After the first charge, it is recommended that you fully
charge the battery periodically, even when it is not in use.
This is because the Ni-MH rechargeable battery pack may
drain with time.
The performance of the rechargeable battery may degrade with time.
NOTE
Charging the battery
Never discharge the battery by shorting it or by reverse polarity in any
WARNING
subjects. Make sure the battery is the rechargeable type before
charging it. Never rotate the rotary switch when the battery is being
charged as DC at 24 V is being applied to the charging terminals.
For the battery charger, the mains supply voltage should not fluctuate by
plus or minus 10%.
NOTE
98
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Maintenance
5
CAUTION
•
•
Do not rotate the rotary switch from
charging the battery.
position when
Perform battery charging only with a 9 V Ni-MH rechargeable
battery (7.2 V nominal voltage) or 9 V size Ni-MH rechargeable
battery (8.4 V nominal voltage).
•
•
Disconnect test leads from all the terminals when charging.
Ensure proper insertion of battery in the multimeter and follow the
correct polarity.
Use the specified 24 V DC adaptor to charge the battery.
Remember never to turn the rotary switch of the meter
when the battery is being charged. Use the following steps to
charge the battery:
1 Disconnect the test leads from the meter.
2 Turn the rotary switch to the
position. Connect
the power cord to the DC adapter.
3 Plug the Red (+)/ Black (–) banana terminals of the DC
adapter to the and the “COM” terminals
respectively. The DC adaptor can be replaced with a DC
power supply in order to set the 24 V DC output and the
over current limitation to a value more than 0.5 A.
Ensure that the polarity of the connection is correct.
4 The primary display will show “bAt” and the ‘SbY” will be
shown on the secondary display and a short beep will
sound to remind you whether you need to charge the
battery. Press SHIFT to start charging the battery, or the
meter will automatically start the self-test after the 24 V
supply is applied. It is recommended that you do not
charge if the battery’s capacity is over 90%.
Agilent U1251A/U1252A User’s and Service Guide
99
Download from Www.Somanuals.com. All Manuals Search And Download.
5
Maintenance
Table 5-1 Battery voltage and corresponding percentage of charges in standby and charging modes
Condition
Trickle (SBY)
Charging
Battery voltage
6.0 V to 8.2 V
7.2 V to 10.0 V
Proportional percentage
0% to 100%
0% to 100%
Figure 5-2 Battery capacity display as trickle
5 After pressing Shift or when the self-test starts, the meter
performs a self-test to check if the battery in the meter is
the rechargeable type. The self-test takes about 2 to 3
minutes. Avoid pressing any button during the self-test. A
message is diplayed as shown below.
Figure 5-3 Self-test
100
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Maintenance
5
Table 5-2 Error messages
Error
Error message
OL
1
2
3
No battery inside
Faulty battery
Battery is fully charged
C-Err
1
2
If charging battery more than 12 V or
less than 5 V
In 3 minutes, if the battery voltage
does not go upwards then charge
error
Agilent U1251A/U1252A User’s and Service Guide
101
Download from Www.Somanuals.com. All Manuals Search And Download.
5
Maintenance
•
•
If the OL message appears while the battery is inside, do not charge the battery.
If the C-Err message appears, check if the battery is the specified type. Use the
battery as specified in this manual. Ensure that the battery is the correct
rechargeable type before charging it again. After replacing with the correct
rechargeable battery, press Shift to perform the self-test again. If the C-Err
message reappears, replace with a new battery.
NOTE
6 The smart charging mode will start if the meter passes
the self-test. The charging time is limited to not more
than 220 minutes. The secondary display will count down
the charging time. During charging, no buttons can be
pressed. The error message may appear during charging to
alert the user of any overcharging of the battery.
Figure 5-4 Charging mode
102
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Maintenance
5
7 The charge end message (C–End) appears on the
secondary display once charging is completed. The trickle
charging current is provided to maintain the battery
capacity. The flashing signs of
show the trickle state.
and
appear to
8 Remove the DC adapter when the C–End message appears
on the secondary display. Do not turn the rotary switch
before removing the adapter from the terminals.
Figure 5-5 Charge end and trickle state
Agilent U1251A/U1252A User’s and Service Guide
103
Download from Www.Somanuals.com. All Manuals Search And Download.
Maintenance
5
Fuse checking procedure
It is recommended that you check the fuses of the
multimeter before using it. Follow the instructions below to
test the fuses inside the multimeter. Refer to Figure 5-8 for
the respective positions of Fuse 1 and Fuse 2.
1 Set the rotary switch to
.
2 Connect the red test lead to the input terminal
.
3 To test Fuse 1, place the tip of the test probe on the right
half of input terminal
. Ensure that the probe tip
touches the metal inside the input terminal, as shown in the
figure above.
Agilent U1251A/U1252A User’s and Service Guide
105
Download from Www.Somanuals.com. All Manuals Search And Download.
5
Maintenance
4 To test Fuse 2, place and touch the tip of the test probe
on the right half of input terminal . Ensure that the
probe tip touches the metal inside the input terminal.
5 Observe the reading on the instrument's display. Refer to
Table 5-3 for the possible readings that could appear).
6 Replace the fuse when OL is displayed.
Table 5-3 Measurement readings for fuse checking
Fuse OK (approximately)
Replace fuse
Current input terminal
Fuse
Fuse rating
Displayed readings
1
2
440 mA/1000 V
11 A/1000 V
6.2 MΩ
0.06 Ω
OL
OL
Replacing the fuse
This manual provides only the fuse replacement procedures, but not the
fuse replacement markings.
NOTE
Follow the procedures below to replace the fuse.
1 Turn the meter off and disconnect the test leads from the
external equipment. Make sure to remove the adaptor.
2 Wear clean/dry gloves and do not touch any other
component except the fuse and plastic parts. The current
calibration is considered shunt only, so do not recalibrate
the meter after replacing the fuse.
3 Remove the battery cover compartment to replace the
fuse.
4 Loosen the three screws on the bottom case and remove
the cover.
5 Loosen the two screws on the top corners to lift the
circuit board.
106
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Maintenance
5
6 Gently remove the defective fuse by prying one end of the
fuse loose and removing it out of the fuse bracket.
7 Replace with a new fuse of the same size and rating.
Make sure the new fuse is centered in the fuse holder.
8 Ensure that the rotary switch on the top case and the
circuit board switch stay on the OFF position.
9 Then re-fasten the circuit board and the bottom cover.
10 Refer to the table below for the part number, rate, and
size of the fuses.
Table 5-4 Fuse specifications
Fuse
Agilent part number
Rating
Size
Type
1
2
2110-1400
2110-1402
440 mA/1000 V
11 A/1000 V
10 mm × 35 mm
10 mm × 38 mm
Fast blow fuse
Fuse 1
Fuse 2
Figure 5-7 Fuse replacement
Agilent U1251A/U1252A User’s and Service Guide
107
Download from Www.Somanuals.com. All Manuals Search And Download.
5
Maintenance
Troubleshooting
To avoid electrical shock, do not perform any service unless you are
qualified to do so.
WARNING
If the instrument fails to operate, check the battery and the
test leads. Replace them if necessary. And if the instrument
still does not function, check the operating procedures in
this manual. When servicing, use only specified replacement
parts. Table 5-5 below will assist you to identify some basic
problems and their solutions.
Table 5-5 Basic troubleshooting procedures
Malfunction
Troubleshooting procedure
Check the battery. Charge or replace it.
•
•
No LCD display after power ON
No beeper tone
Check the setup mode to verify if the beeper is set to OFF. Then select the
desired driving frequency.
•
Check the fuse.
Fail on current measurement
•
•
Check 440 mA fuse.
[1]
No charging indication
Check the external adaptor to verify if the output is DC 24 V and the adaptor is
plugged into the charging terminals completely.
Line power voltage (100V~240 VAC 50 Hz/ 60 Hz).
•
•
•
Check if the correct rechargeable battery is used.
Try to charge and discharge for 2 or 3 cycles in order to maintain the battery’s
highest capacity.
Battery life very short after full
charge/Battery not able to charge
after pro-long storage period
• NOTE: The performance of the rechargeable battery may degrade with time.
•
Check that the optical side of the cable is connected to the meter and the text
side of the cover is up.
Fail on Remote control
•
•
Check the baud rate, parity, Data bit, Stop bit (default is 9600, n, 8, 1)
Driver install for IR-USB.
Notes for basic troubleshooting procedures table:
Never turn the rotary switch of the multimeter from the OFF position when it is charging.
1
108
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Maintenance
5
Replaceable Parts
This section contains information for ordering replacement
parts for your instrument. you can find the instrument
support part list at Agilent’s Test & Measurement Parts
Catalog at: http://www.agilent.com/find/parts
The parts lists include a brief description of each part with
applicable Agilent part number.
To order replaceable parts
You can order replaceable parts from Agilent using the
Agilent part number. Note that not all parts listed are
available as field-replaceable parts.
To order replaceable parts from Agilent, do the following:
1 Contact your nearest Agilent Sales Office or Service
Center.
2 Identify the parts by the Agilent part number shown in
the support parts list.
3 Provide the instrument model number and serial number.
Agilent U1251A/U1252A User’s and Service Guide
109
Download from Www.Somanuals.com. All Manuals Search And Download.
5
Maintenance
110
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent U1251A and U1252A Handheld Digital Multimeter
User’s and Service Guide
6
Adjustments Consideration 133
Valid adjustment input values 134
Adjustment procedure 135
Finishing the adjustment 142
Calibration errors 143
This chapter contains the performance test and the
adjustment procedure.
Agilent Technologies
111
Download from Www.Somanuals.com. All Manuals Search And Download.
7
Performance Tests and Calibration
Calibration Overview
This manual contains procedures to verify the instrument’s
performance and adjustment (calibration).
The performance test procedures allow you to verify that the
handheld digital multimeter is operating within its published
specifications. The adjustment procedures ensure that the
multimeter remains within its specifications until the next
calibration.
Make sure you have read the “Test Considerations” on page 118 before
calibrating the instrument.
NOTE
Closed-case electronic calibration
This instrument features closed-case electronic calibration.
No internal mechanical adjustments are required. The
instrument calculates correction factors based upon the
input reference value you set. The new correction factors are
stored in the nonvolatile memory until the next calibration
adjustment is performed. The nonvolatile EEPROM
calibration memory is retained even when the power is
switched off.
Agilent Technologies calibration services
When your instrument is due for calibration, contact your
local Agilent Service Center to enquire about recalibration
services.
112
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Performance Tests and Calibration
7
Calibration interval
A 1-year interval is adequate for most applications. Accuracy
specifications are warranted only if adjustment is made at
regular calibration intervals. Accuracy specifications are not
warranted beyond the 1-year calibration interval. Agilent
does not recommend extending calibration intervals beyond
2 years for any application.
Adjustment is recommended
Specifications are only guaranteed within the period stated
from the last adjustment. Agilent recommends that
readjustment should be performed during the calibration
process for best performance. This will ensure that the
U1251A/U1252A will remain within the specifications for the
next calibration interval. This criterion for the re-adjustment
provides the best long–term stability.
Performance data are measured during the Performance
instrument will remain within these limits unless the
adjustments are performed.
Refer to the section “To read the calibration count” on
page 142 and verify that all the adjustments have been
performed.
Agilent U1251A/U1252A User’s and Service Guide
113
Download from Www.Somanuals.com. All Manuals Search And Download.
7
Performance Tests and Calibration
Recommended Test Equipment
The test equipment recommended for the performance
verification and adjustment procedures is listed below. If the
exact instrument is not available, substitute calibration
standards of equivalent accuracy.
A suggested alternative method would be to use the Agilent
3458A 8½ – Digit Digital Multimeter to measure less
accurate yet stable sources. The output value measured from
the source can be entered into the instrument as the target
calibration value.
Table 6-1 Recommended Test Equipment
Application
DC Voltage
DC Current
Resistance
AC Voltage
AC Current
Frequency
Capacitance
Duty Cycle
Nanosiemens
Diode
Recommended equipment
Fluke 5520A
Recommended accuracy requirements
<1/5 instrument 1 year spec
<1/5 instrument 1 year spec
<1/5 instrument 1 year spec
<1/5 instrument 1 year spec
<1/5 instrument 1 year spec
<1/5 instrument 1 year spec
<1/5 instrument 1 year spec
<1/5 instrument 1 year spec
<1/5 instrument 1 year spec
<1/5 instrument 1 year spec
Fluke 5520A
Fluke 5520A
Fluke 5520A
Fluke 5520A
Agilent 33250A
Fluke 5520A
Fluke 5520A
Fluke 5520A
Fluke 5520A
Frequency
Counter
Agilent 33250A
<1/5 instrument 1 year spec
Temperature
Square Wave
Fluke 5520A
<1/5 instrument 1 year spec
<1/5 instrument 1 year spec
Agilent 53131A and Agilent 34401A
Shorting Plug - Dual banana plug with copper wire
short between 2 terminal
Short
–
114
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Performance Tests and Calibration
7
Basic Operating Test
The Basic Operating Test is to test the basic operability of
the instrument. Repair is required if the instrument fails the
Basic Operating Test.
Backlit test
Press the Bat button to test the backlight. It will
momentarily toggle backlit ON and OFF.
Testing the display
Press the Hold button and turn on the meter to view all
segments of the display. Compare the display with the
example in Figure 6-1.
Figure 6-1 LCD display
Agilent U1251A/U1252A User’s and Service Guide
115
Download from Www.Somanuals.com. All Manuals Search And Download.
7
Performance Tests and Calibration
Current terminal test
This test determines if the input warning of the current
The meter sounds an alert beep when the test lead is
inserted into the A terminal but the rotary switch is not set
to mA.A function. The primary display will indicate “A-Err”
as shown in Figure 6-2. The primary display will keep flash-
ing unless the test lead is removed from “A” terminal.
Before conducting this test, make sure the beep function is not disabled in
setup.
NOTE
Figure 6-2 Input warning
116
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Performance Tests and Calibration
7
Charge terminal alert test
This test determines if the charge terminal alert is operating
properly.
The meter sounds an alert when the
terminal
detects a voltage level of more than 5 V and the rotary
switch is not set to the
position. The meter sounds
an alert beep and the primary display flashes the “Ch.Err”
until the lead is removed from the
terminal.
Figure 6-3 Charge terminal alert
Before conducting this test, make sure the beep function is not disabled at
setup.
NOTE
Agilent U1251A/U1252A User’s and Service Guide
117
Download from Www.Somanuals.com. All Manuals Search And Download.
7
Performance Tests and Calibration
Test Considerations
Long test leads can also act as antennas which may pick up
AC signals.
For optimum performance, all procedures should comply
with the following recommendations:
• Ensure that the calibration ambient temperature is stable
and is between 18 °C and 28 °C. Ideally the calibration
should be performed at 23 °C 1 °C.
• Ensure that the ambient relative humidity is less than
80%.
• Allow a warm-up period of five minutes.
• Use shielded twisted pair PTFE-insulated cables to reduce
settling and noise errors. Keep the input cables as short
as possible.
118
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Performance Tests and Calibration
7
Calibration Security
The calibration security code prevents accidental or
unauthorized adjustments to the instrument. When you first
receive your instrument, it is secured. Before you can adjust
the instrument, you must unsecure it by entering the correct
security code (see “Unsecuring the instrument for
calibration” on page 128).
The security code is set to 1234 when the instrument is
shipped from the factory. This code is stored in nonvolatile
memory, and does not change when power is turned off.
You can unsecure the instrument from the front panel. The security code
can only be changed from the front panel after the instrument has been
unsecured.
NOTE
See “To unsecure the instrument without the security code” on page 130 if
you forget your security code.
Agilent U1251A/U1252A User’s and Service Guide
119
Download from Www.Somanuals.com. All Manuals Search And Download.
7
Performance Tests and Calibration
Performance Verification Tests
Use the Performance Verification Tests to verify the
measurement performance of the instrument. The
performance verification tests use the instrument’s
specifications listed in the U1251A/U1252A Data Sheet.
The performance verification tests are recommended as
acceptance tests when you first receive the instrument. The
acceptance test results should be compared against the one
year test limits. After acceptance, you should repeat the
performance verification tests at every calibration interval.
Make sure you have read the “Test Considerations” on page 118 before
doing the performance verification tests.
NOTE
Perform the verification test steps in the following Table 6-2
120
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Performance Tests and Calibration
7
Table 6-2 Performance verification test steps
Error from nominal 1 year
Step
Test Function
Range
5520A Output
U1251A
32.5 mV
U1252A
22.5 mV
1
Turn the rotary switch to the
5 V
5 V, 1 kHz
[1]
V position
5 V, 10 kHz
5 V, 20 kHz
5 V, 30 kHz
52.5 mV
N/A
84 mV
22.5 mV
41.5 mV
N/A
4.5 V, 100 kHz N/A
169.5 mV
225 mV
225 mV
415 mV
50 V
50 V,1 kHz
325 mV
525 mV
N/A
840 mV
N/A
50 V,10 kHz
50 V, 20 kHz
50 V, 30 kHz
45 V, 100 kHz
500 V, 1 kHz
500 V, 10 kHz
1000 V, 1 kHz
0.48 V, 1 kHz
N/A
1.695 V
2.25 V
2.25V
500 V
3.25 V
5.25 V
10 V
1000 V
8.0 V
2
3
Press
frequency mode
button to go to
9.9999 kHz
500 mHz
500 mHz
Press
button to go to
0.01% – 99.99%
5.0 Vpp @
0.315%
0.315%
Duty Cycle mode
50%, Square
Wave, 50 Hz
Agilent U1251A/U1252A User’s and Service Guide
121
Download from Www.Somanuals.com. All Manuals Search And Download.
7
Performance Tests and Calibration
Error from nominal 1 year
Step
Test Function
Range
5520A Output
U1251A
2 mV
U1252A
1.75 mV
4
Turn the rotary switch to
5 V
5 V
V position (for model U1252A), to
V position (for model
U1251A)
50 V
50 V
20 mV
17.5 mV
200 mV
800 mV
22.5 mV
500 V
1000 V
5 V
500 V
200 mV
800 mV
1000 V
5 V,1 kHz
5
Press
button to go to the
V mode
N/A
[1]
5 V, 10 kHz
5 V, 20 kHz
N/A
N/A
22.5 mV
41.5 mV
4.5 V, 100 kHz N/A
169.5 mV
225 mV
225 mV
415 mV
1.695 V
2.25 V
50 V
50 V, 1 kHz
N/A
N/A
N/A
N/A
N/A
N/A
N/A
50 V, 10 kHz
50 V, 20 kHz
45 V, 100 kHz
500 V, 1 kHz
500 V, 10 kHz
1000 V, 1 kHz
500 V
2.25 V
1000 V
8.0 V
122
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Performance Tests and Calibration
Error from nominal 1 year
7
Step
Test Function
Range
5520A Output
U1251A
U1252A
[2]
[2]
6
Turn the rotary switch to the
mV position
50 mV
50 mV
75 μV
75 μV
500 mV
1000 mV
50 mV
500 mV
0.2 mV
0.2 mV
0.8 mV
0.8 mV
0.34 mV
0.175 mV
0.175 mV
0.75 mV
0.75 mV
0.24 mV
– 500 mV
1000 mV
– 1000 mV
50 mV, 1 kHz
7
Press
the
button to go to
mV mode
[1]
50 mV, 10 kHz
50 mV, 20 kHz
50 mV, 30 kHz
45 mV, 100 kHz
500 mV, 45 Hz
500 mV, 1 kHz
500 mV, 10 kHz
500 mV, 20 kHz
500 mV, 30 kHz
450 mV, 100 kHz
1000 mV, 1 kHz
1000 mV, 10 kHz
1000 mV, 20 kHz
1000 mV, 30 kHz
1000 mV, 100 kHz
0.54 mV
0.39 mV
N/A
0.86 mV
N/A
0.415 mV
N/A
1.695 mV
2.25 mV
2.25 mV
2.25 mV
4.15 mV
500 mV
3.25 mV
3.25 mV
5.4 mV
N/A
8.6 mV
N/A
N/A
16.95 mV
6.5 mV
1000 mV
8.5 mV
12.5 mV
6.5 mV
N/A
20.0 mV
N/A
11.5 mV
N/A
47 mV
Agilent U1251A/U1252A User’s and Service Guide
123
Download from Www.Somanuals.com. All Manuals Search And Download.
7
Performance Tests and Calibration
Step
Test Function
Range
5520A Output
U1251A
U1252A
[3]
[3]
8
Turn the rotary switch to the W
500 W
500 W
500 mW
350 mW
position
[3]
[3]
5 kW
5 kW
4.5 W
45 W
3 W
50 kW
500 kW
5 MW
50 MW
50 kW
5 MW
30 W
450 W
300 W
10.5 kW
8 kW
[4]
50 MW
450 MW
50 nS
0.510 MW
0.505 MW
36.1 MW
0.6 nS
500 MW
N/A
[5]
9
Press
mode
button to go to ns
500 nS
0.7 nS
10
Turn the rotary switch to the Hz/
position (for model
Diode
1 V
1 mV
1 mV
U1252A), to
position (for
model U1251A)
33250A Output
11
12
Press
button to go to
999.99 kHz
99.999 MHz
200 mVrms,
100 kHz
N/A
52 Hz
[6]
frequency counter mode
Press
button to go to
600 mVrms,
10 MHz
N/A
5.2 kHz
frequency counter mode divide
by 100
5520A Output
13
Turn the rotary switch to the
10.000 nF
10.000 nF
0.108 nF
0.108 nF
[7]
TEMP/
position
100.00 nF
1000.0 nF
10.000 μF
100.00 nF
1000.0 nF
10.000 μF
1.05 nF
10.5 nF
0.105 μF
1.05 nF
10.5 nF
0.105 μF
124
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Performance Tests and Calibration
Error from nominal 1 year
7
Step
Test Function
Range
5520A Output
U1251A
1.05 μF
U1252A
1.05 μF
100.00 μF
1000.0 μF
10.00 mF
100.00 mF
100.00 μF
1000.0 μF
10.00 mF
10.00 mF
0 ºC
10.5 μF
0.105 mF
0.4 mF
3 ºC
10.5 μF
0.105 mF
0.4 mF
3 ºC
14
15
16
Press
TEMP mode
button to go to
–200 ºC until
1372 ºC
[8][13]
100 ºC
3.3 ºC
3.3 ºC
[9]
[9]
Turn the rotary swich to the
uA position
500 μA
500 μA
0.55 μA
0.3 μA
[9]
[9]
5000 μA
500 μA
5000 μA
5.5 μA
4.2 μA
3 μA
Press
button to go to
500 μA, 1 kHz
3.7 μA
[1]
uA mode
500 μA, 20 kHz
5000 μA, 1 kHz
5000 μA, 20 kHz
50 mA
15.8 μA
42 μA
3.95 μA
39.5 μA
5000 μA
0.156 mA
0.105 mA
[9]
[9]
17
18
Turn the rotary switch to the
mA.A position
50 mA
80 μA
[9]
[9]
440 mA
50 mA
400 mA
0.93 mA
0.42 mA
0.71 mA
0.37 mA
Press
button to go to
50 mA, 1 kHz
[1]
mA mode
50 mA, 20 kHz
400 mA, 45 Hz
400 mA, 1 kHz
1.56 mA
3.4 mA
3.4 mA
0.395 mA
3.0 mA
440 mA
3.0 mA
Caution: Connect calibrator to hanheld multimeter's A and COM terminal before applying 5A and 10A
5 A
5 A
16 mA
40 mA
16 mA
35 mA
[10]
10 A
10 A
Agilent U1251A/U1252A User’s and Service Guide
125
Download from Www.Somanuals.com. All Manuals Search And Download.
7
Performance Tests and Calibration
Error from nominal 1 year
Step
Test Function
Range
5520A Output
U1251A
42 mA
U1252A
37 mA
19
Press
button to go to
A mode
5 A
5 A, 1 kHz
3A
3 A, 5 kHz
96 mA
96 mA
90 mA
[11]
10 A
10 A, 1 kHz
Use 53131A
100 mA
Square Wave
Output
20
Turn the rotary switch to the
position
120 Hz @ 50%
N/A
26 mHz
4800 Hz @ 50%
100 Hz @ 50%
100 Hz @ 25%
100 Hz @ 75%
N/A
N/A
N/A
N/A
260 mHz
[12]
Duty Cycle
0.398%
0.398%
0.398%
[12]
[12]
Use 34410A
Amplitude
4800 Hz @
99.609%
N/A
0.2V
Notes for verification test:
1
2
The additional error to be added as frequency > 20 kHz and signal input < 10% of range: 300 counts of LSD per kHz.
The accuracy could be 0.05% + 10, always use relative function to zero thermal effect (short test leads) before measuring
the signal.
3
4
5
6
The accuracy of 500 W and 5 kW is specified after the Null function.
For range of 50 MW, the RH is specified for < 60%.
The accuracy is specified for < 50 nS and after Null function as open test lead.
All frequency counters are susceptible to error when measuring low voltage, low-frequency signals. Shielding inputs from
external noise pickup is critical for minimizing measurement errors.
7
8
Use the Null mode to zero residual.
The accuracy does not include the tolerance of the thermocouple probe. The thermal sensor plugged into the meter should
be placed in the operating enviroment for at least an hour. Use the Null function to reduce the thermal effect.
Always use relative function to zero the thermal effect with open test lead before measuring the signal. If you do not use
Relation function, add 20 digits for accuracy purposes.
9
10 10 A continuous and additional 0.5% to specified accuracy when measuring a signal greater than 10 A~20 A for 30 seconds
maximum. After measuring current for > 10 A, to cool down the meter for twice the measuring time you applied before low
current measurement.
126
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Performance Tests and Calibration
7
11 The current can be measuring from 2.5 A to 10 A continuous, and the additional of 0.5% to specified accuracy as measure
the signal greater than 10 A~20 A for 30 seconds maximum. After measured current for > 10 A, to cool down the meter for
2 times of measuring time you applied before low current measurement.
12 For signal frequency greater than 1 kHz, additional 0.1% per kHz to be added to accuracy.
13 Ensure that the ambient temperature is stable within 1ºC. Make sure that the multimeter is placed in a controlled
environment for at least 1 hour. Keep the multimeter away from any ventilation exit. Do not touch the thermocouple test
lead after connecting it to the calibrator. Allow the connection to stabilize for at least another 15 minutes before performing
the measurement.
Agilent U1251A/U1252A User’s and Service Guide
127
Download from Www.Somanuals.com. All Manuals Search And Download.
7
Performance Tests and Calibration
Unsecuring the instrument for calibration
Before you can adjust the instrument, you must unsecure it
by entering the correct security code. The security code is
set to 1234 when the instrument is shipped from the
factory. The security code is stored in non-volatile memory,
and does not change when power is off.
To Unsecure the Instrument from the Front Panel
4 Turn the rotary switch to
.
5 Press
and
button simultaneously to enter
the Calibration Security Code entry mode.
6 The primary display shows 5555 and the secondary
display show SECUr.
7 Use the editing keys
and
to step each
character in the code.
Use the
and
to select each character.
8 Press
(Save) when done.
9 If the correct security code is entered, the secondary
display will show PASS.
128
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Performance Tests and Calibration
7
To change the instrument calibration security code from the front
panel
1 When the multimeter is in the unsecured mode, press
for more than 1 second to enter the Calibration
Security Code setting mode.
2 The factory default calibration security code 1234 will be
displayed on the primary display.
3 Use the editing keys
and
to step each
character in the code.
4 Use the
and
to change each character in
the code.
5 Press
(Save) to store the new calibration security
code.
6 If the new calibration security code has been successfully
stored, the secondary display will show PASS.
Agilent U1251A/U1252A User’s and Service Guide
129
Download from Www.Somanuals.com. All Manuals Search And Download.
7
Performance Tests and Calibration
To unsecure the instrument without the security code
To unsecure the instrument without the correct security
code, follow the steps below.
If you do not have a record of the security code, you can try 1234 (the fac-
tory default code) using the front panel.
NOTE
1 Record the last 4 digits of the multimeter’s serial number.
2 Turn the rotary switch to
.
3 Press and
simultaneously to enter the
Calibration Security Code entry mode.
4 The primary display shows 5555 and the secondary
display shows SECUr.
5 Press
for more than 1 second to enter the Set
Default Security Code mode. The secondary display shows
SEr.no and the primary display shows 5555.
6 Use the editing keys
and
to step each
character in the code.
Use
and
to select each character.
7 Set the code, same as the last 4 digits of the instrument’s
serial number.
8 Press
(Save) to confirm the entry.
9 If the 4 digits entered are correct, the secondary display
will show PASS.
130
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Now you can use 1234 as the security code. If you want to
enter a new security code, see “To change the instrument
calibration security code from the front panel” on page 129.
Make sure you record the new security code.
Calibration process
To complete a full instrument calibration use following
general procedures:
1 Read “Test Considerations” on page 118.
instrument (incoming data).
3 Unsecure the instrument for calibration (see “Calibration
Security” on page 119).
4 Perform the adjustment procedures (see “Adjustments
Consideration” on page 133).
5 Secure the instrument against calibration.
6 Note the new security code and calibration count in the
instrument's maintenance records.
Make sure to quit the Adjustment Mode before turning off the instrument.
NOTE
Agilent U1251A/U1252A User’s and Service Guide
131
Download from Www.Somanuals.com. All Manuals Search And Download.
7
Performance Tests and Calibration
Using the front panel for adjustments
Selecting the Adjustment Mode
To unsecure the instrument, see “Unsecuring the instrument
for calibration” on page 128 or “To unsecure the instrument
without the security code” on page 130. Once unsecured, the
reference value will be indicated on the primary display.
Entering Adjustment Values
Use the following adjustment procedure to enter an input
calibration value from the front panel:
1 Use the edit keys
and
to select each digit in
the primary display.
2 Use the
through the digits 0 through 9.
3 Press when done to start calibration.
and the
arrow keys to advance
132
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Performance Tests and Calibration
7
Adjustments Consideration
Shorting Plug to adjust the instrument.
After each adjustment, the secondary display briefly shows PASS. If the
calibration fails, the handheld multimeter sounds a beep, and an error
NOTE
number is shown in the secondary display. Calibration error messages are
described on page 143. In the event of a calibration failure, correct the
problem and repeat the procedure.
Adjustments for each function should be performed only in
the order shown below.
1 Allow the instrument to warm up and stabilize for 5
minutes before performing the adjustments.
2 Ensure that during the adjustments, the low battery
indicator does not appear. Replace the batteries as soon
as possible to avoid false reading.
3 Consider the thermal effects as you are connecting test
leads to the calibrator and handheld multimeter. It is
recommended to wait for one minute before you begin the
calibration after connecting the test leads.
4 During ambient temperature adjustment, be sure to turn
on the instrument for at least 1 hour with the K-type
thermocouple connected between the instrument and the
calibration source.
Never turn off the instrument during an adjustment. This may delete the
calibration memory for the present function.
CAUTION
Agilent U1251A/U1252A User’s and Service Guide
133
Download from Www.Somanuals.com. All Manuals Search And Download.
7
Performance Tests and Calibration
Valid adjustment input values
Adjustment can be accomplished using the following input
values below.
Table 6-3 Valid adjustment input values
Function
Range
Valid Amplitude Input Values
0.9 to 1.1 x Full Scale
5 V, 50 V, 500 V, 1000 V
0.9 to 1.1 x Full Scale
5 V, 50 V, 500 V, 1000 V
V
(for U1251A)
0.9 to 1.1 x Full Scale
0.9 to 1.1 x Full Scale
0.9 to 1.1 x Full Scale
0.9 to 1.1 x Full Scale
0.9 to 1.1 x Full Scale
5 V, 50 V ,500 V, 1000 V
50 mV, 500 mV, 1000 mV
500 µA, 5000 µA
(for U1252A)
50 mA, 440 mA, 5 A, 10 A
Ω
500 Ω, 5 kΩ, 50 kΩ,500 kΩ, 5 MΩ,
50 MΩ
/ TEMP
10 nF, 100 nF, 1000 nF, 10 µF, 100 µF,
Make sure to provide 0 °C with ambient
1000 µF, 10 mF / 0 °C
compensation
134
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Adjustment procedure
Review the sections “Test Considerations” on page 118 and
“Adjustments Consideration” on page 133 before proceeding
with this procedure.
1 Turn the rotary switch to “Test Function” position as
shown in the adjustment table.
2 After unsecuring the instrument, the instrument goes into
the adjustment mode. (See “Unsecuring the instrument for
calibration” on page 128)
The instrument will be in the adjustment mode, unless you Press SHIFT
NOTE
and
simultaneously to exit the adjustment mode.
3 The primary display will show the reference value of the
Cal Item.
4 Configure each Cal Item.
5 Use the
Cal Range.
and the
arrow keys to select the
6 Apply the input signal shown in the Input column of the
table. The bar graph will display the Input reading. There
is no bar graph display for temperature adjustment.
Always complete the tests in the same order as shown in the appropriate
table.
NOTE
7 Enter the actual applied input (see “Entering Adjustment
Values” on page 132).
Agilent U1251A/U1252A User’s and Service Guide
135
Download from Www.Somanuals.com. All Manuals Search And Download.
7
Performance Tests and Calibration
8 Press
to start the adjustment. The CAL flashes in
the secondary display to indicate that the calibration is in
progress.
For an adjustment that is successful, the secondary
display will briefly show PASS. An adjustment failure is
indicated by a long beep and a calibration error number
appears in the secondary display. The primary display
range, function, and entered adjustment value to fix the
problem. Repeat the adjustment procedure.
9 Repeat steps 1 through 8 for each adjustment point.
10 Verify the adjustments using the “Performance Verification
Tests” on page 120 and Table 6-2.
136
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Performance Tests and Calibration
7
Table 6-4 Adjustment table
Cal Item
Step
Test Function
Cal Range
Input
U1251A
U1252A
1
Turn the rotary switch to the
V position
5 V
0.3 V,1 kHz
0.3000 V
0.3000 V
3 V, 1 kHz
3.0000 V
N/A
3.0000 V
3.0000 V
03.000 V
30.000 V
30.000 V
030.00 V
300.00 V
300.00 V
0030.0 V
0300.0 V
0300.0 V
SHort
3 V, 50 kHz
3 V, 1 kHz
50 V
03.000 V
30.000 V
N/A
30 V, 1 kHz
30 V, 50 kHz
30 V,1 kHz
300 V,1 kHz
300 V, 50 kHz
30 V, 1 kHz
300 V, 1 kHz
300 V, 50 kHz
500 V
1000 V
Short
030.00 V
300.00 V
N/A
0030.0 V
0300.0 V
N/A
2
Turn the rotary switch to
Dual Banana Plug
with copper wire
short between 2
terminals
SHort
V position (for model U1252A),
to
V position (for model
U1251A)
5 V
3 V
3.0000 V
30.000 V
300.00 V
1000.0 V
3.0000 V
30.000 V
300.00 V
1000.0 V
50 V
30 V
500 V
1000 V
300 V
1000 V
Agilent U1251A/U1252A User’s and Service Guide
137
Download from Www.Somanuals.com. All Manuals Search And Download.
7
Performance Tests and Calibration
Cal Item
U1252A
Step
Test Function
Cal Range
Input
U1251A
3
Press
button to go to the
V mode
5 V
0.3 V, 1 kHz
N/A
0.3000 V
3 V, 1 kHz
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
SHort
3.0000 V
3.0000 V
03.000 V
30.000 V
30.000 V
030.00 V
300.00 V
300.00 V
0030.0 V
0300.0 V
0300.0 V
SHort
3 V, 50 kHz
3 V, 1 kHz
50 V
30 V, 1 kHz
30 V, 50 kHz
30 V, 1 kHz
300 V, 1 kHz
300 V, 50 kHz
30 V, 1 kHz
300 V, 1 kHz
300 V, 50 kHz
500 V
1000 V
Short
4
Turn the rotary switch to the
mV position
Dual banana plug
with copper wire
short between 2
terminals
50 mV
30 mV
30.000 mV
300.00 mV
1000.0 mV
30.000 mV
300.00 mV
1000.0 mV
500 mV
1000 mV
300 mV
1000 mV
138
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Performance Tests and Calibration
7
Cal Item
Step
Test Function
Cal Range
Input
U1251A
U1252A
5
Press
button to go to the
mV mode
50 mV
3 mV, 1 kHz
03.000 mV
03.000 mV
30 mV, 1 kHz
30.000 mV
N/A
30.000 mV
30.000 mV
030.00 mV
300.00 mV
300.00 mV
0030.0 mV
1000.0 mV
1000.0 mV
SHort
30 mV, 50 kHz
30 mV, 1 kHz
500 mV
1000 mV
Short
030.00 mV
300.00 mV
N/A
300 mV, 1 kHz
300 mV, 50 kHz
30 mV, 1 kHz
0030.0 mV
1000.0 mV
N/A
1000 mV, 1 kHz
1000 mV, 50 kHz
6
Turn the rotary switch to the W
Dual Banana
Plug with copper
wire short
SHort
position
between 2
terminals
50 MW
Input terminal
open (Remove
any test leads
and Short Plugs
from the input
terminal)
oPEn
oPEn
10 MW
3 MW
300 kW
30 kW
3k W
10.000 MW
3.0000 MW
300.00 kW
30.000 kW
3.0000 kW
300.00 W
10.000 MW
3.0000 MW
300.00 kW
30.000 kW
3.0000 kW
300.00 W
5 MW
500 kW
50 kW
5 kW
500 W
300 W
Agilent U1251A/U1252A User’s and Service Guide
139
Download from Www.Somanuals.com. All Manuals Search And Download.
7
Performance Tests and Calibration
Cal Item
U1252A
Step
Test Function
Cal Range
Input
U1251A
7
Turn the rotary switch to the
Open
Input terminal
oPEn
oPEn
TEMP/
position
open (Remove any
test leads and
Short Plugs from
the input terminal)
10 nF
3 nF
03.000 nF
10.000 nF
010.00 nF
100.00 nF
0100.0 nF
1000.0 nF
10.000 μF
100.00 μF
1000.0 μF
10.000 mF
0000.0 ºC
03.000 nF
10.000 nF
010.00 nF
100.00 nF
0100.0 nF
1000.0 nF
10.000 μF
100.00 μF
1000.0 μF
10.000 mF
0000.0 ºC
10 nF
100 nF
1000 nF
10 nF
100 nF
100 nF
1000 nF
10 μF
100 μF
1000 μF
10 mF
0 ºC
10 μF
100 μF
1000 μF
10 mF
N/A
8
9
Press
TEMP mode
button to go to
Note: Ensure the multimeter is turned on and stabilized for at least 60 minutes, with the K-type thermocouple
connected between the multimeter and the calibrator output terminal
Turn the rotary swich to the
OPEN
Input terminal
oPEn
oPEn
mA position
open (Remove any
test leads and
Short Plugs from
the input terminal)
500 μA
300 μA
300.00 μA
3000.0 μA
300.00 μA
3000.0 μA
5000 μA
3000 μA
140
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Performance Tests and Calibration
7
10
11
Press
button to go to
mA mode
500 μA
5000 μA
Open
30 μA,1 kHz
030.00 μA
300.00 μA
0300.0 μA
3000.0 μA
oPEn
030.00 μA
300.00 μA
0300.0 μA
3000.0 μA
oPEn
300 μA,1 kHz
300 μA,1 kHz
3000 μA, 1 kHz
Turn the Rotary Switch to the
mA.A position
Input terminal
open (Remove any
test leads and
Short Plugs from
the input terminal)
50 mA
30 mA
30.000 mA
300.00 mA
30.000 mA
300.00 mA
440 mA
300 mA
Move the test lead from uA.mA and COM terminal to A and COM terminal
Caution: Connect calibrator to hanheld multimeter's A and COM terminal before applying 3A and 10A
5 A
3 A
3.0000 A
10.000 A
3.0000 A
10.000 A
10 A
10 A
Move the test lead from A and COM terminal to uA.mA and COM terminal
12
Press
button to go to
mA mode
50 mA
3 mA, 1 kHz
03.000 mA
03.000 mA
30 mA, 1 kHz
30 mA, 1 kHz
300 mA, 1 kHz
30.000 mA
030.00 mA
300.00 mA
30.000 mA
030.00 mA
300.00 mA
440 mA
Move the test lead from uA.mA and COM terminal to A and COM terminal
Caution: Connect calibrator to hanheld multimeter's A and COM terminal before applying 3A and 10A
13
Press
button to go to
A mode
5 A
0.3 A, 1 kHz
0.3000 A
0.3000 A
3 A, 1 kHz
3 A, 1 kHz
10 A, 1 kHz
3.0000 A
3.0000 A
10.000 A
3.0000 A
3.0000 A
10.000 A
10 A
Agilent U1251A/U1252A User’s and Service Guide
141
Download from Www.Somanuals.com. All Manuals Search And Download.
7
Performance Tests and Calibration
Finishing the adjustment
1 Remove all the shorting plugs and connectors from the
instrument.
2 Record the new Calibration Count.
3 Press
and
simultaneously to exit the
Adjustment Mode. Power off and on again. The instrument
will then be secured.
To read the calibration count
You can query the instrument to determine how many
calibrations have been performed.
Your instrument was calibrated before it left the factory.
NOTE
When you receive your instrument, read the count to
determine its initial value.
The count value increases by one for each calibration point,
and a complete calibration will increase the value by many
counts. The calibration count increments up to a maximum
of 65535 after which it is set at 0. The calibration count can
be read from the front panel after the instrument has been
unsecured. Use the following procedure to read the
calibration count from the front panel.
1 Press
shows the calibration count.
2 Take note of the count.
Adjustment Mode. The primary display
3 Press
again to exit the calibration count mode.
142
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Performance Tests and Calibration
7
Calibration errors
The following errors indicate failures that may occur during
a calibration.
Number
200
Recommended Accuracy Requirements
Calibration error: calibration mode is secured
Calibration error: secure code invalid
Calibration error: serial number code invalid
Calibration error: calibration aborted
Calibration error: value out of range
002
003
004
005
006
Calibration error: signal measurement out of range
Calibration error: frequency out of range
EEPROM write failure
007
008
Agilent U1251A/U1252A User’s and Service Guide
143
Download from Www.Somanuals.com. All Manuals Search And Download.
7
Performance Tests and Calibration
144
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Agilent U1251A and U1252A Handheld Digital Multimeter
User’s and Service Guide
7
Frequency Counter Specifications for U1252A 160
Square Wave Output for U1252A 161
Operating Specifications 162
Display update rate (approximate) 162
Input impedance 163
This chapter lists the product characteristics, specification
assumptions and the specifications of the U1251A and
U1252A digital multimeters.
Agilent Technologies
145
Download from Www.Somanuals.com. All Manuals Search And Download.
8
Specifications
Product Characteristics
POWER SUPPLY
Battery type:
•
•
•
•
9 V size Ni-MH rechargeable battery, 7.2 V nominal voltage
9 V size Ni-MH rechargeable battery, 8.4 V nominal voltage
9 V Alkaline battery (ANSI/NEDA 1604A or IEC 6LR61)
9 V Carbon-zinc battery (ANSI/NEDA 1604D or IEC6F22)
Battery life:
•
8 hours typical (based on a fully charged Ni-MH 300 mAH battery for DC
voltage measurement)
•
14 hours typical (based on a new 9 V Alkaline battery for DC voltage
measurement)
Charge time (U1252A only):
•
Less than 220 minutes in an environment of 10 °C to 30 °C. if the battery has
been deep-discharged, a prolonged charging time is required in order for the
battery to return to full capacity
POWER CONSUMPTION
•
•
105 mVA / 420 mVA (with backlight usage) maximum (U1251A)
165 mVA / 480 mVA (with backlight usage) maximum (U1252A)
DISPLAY
•
Both primary and secondary displays are 5-digit liquit crystal display (LCD)
with maximum reading of 50,000 counts
Automatic polarity indication
•
OPERATING ENVIRONMENT
•
•
Temperature: Full accuracy from –20 °C to 55 °C.
Humidity: Full accuracy up to 80% RH (relative humidity) for temperatures up
to 35°C, decreasing linearly to 50% RH at 55 °C.
Altitude:
•
•
•
0 to 2000 meters complying with IEC 61010-1 2nd Edition CAT III, 1000 V/
CAT IV, 600 V.
Pollution degree II
STORAGE COMPLIANCE
–40 °C to 70 °C with battery removed
146
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Specifications
8
SAFETY COMPLIANCE
•
•
•
EN/IEC 61010-1:2001
ANSI/UL 61010-1:2004
CAN/CSA-C22.2 No. 61010-1-04
MEASUREMENT CATEGORY
CAT III 1000 V/CAT IV 600 V Overvoltage Protection
EMC COMPLIANCE
•
•
•
•
Certified to IEC61326-1:2005 / EN61326-1:2006
CISPR 11:2003 / EN 55011:2007 Group 1 Class A
Canada: ICES-001:2004
Austrailia/New Zealand; AS/NZS CISPR11:2004
SHOCK AND VIBRATION
Tested to IEC/EN 60068-2
TEMPERATURE COEFFICIENT
0.15 × (specified accuracy) / °C (from –20 °C to 18 °C, or 28 °C to 55 °C)
COMMON MODE REJECTION RATIO (CMRR)
> 90 dB at DC, 50/60 Hz 0.1% (1 kΩ unbalanced)
NORMAL MODE REJECTION RATION (NMRR)
> 60 dB at 50/60 Hz 0.1%
DIMENSIONS (W × H × D)
94.4 × 203.5 × 59 mm
WEIGHT
•
•
504 5 grams with battery (U1251A)
527 5 grams with battery (U1252A)
Agilent U1251A/U1252A User’s and Service Guide
147
Download from Www.Somanuals.com. All Manuals Search And Download.
8
Specifications
WARRANTY
Please refer to http://www.agilent.com/go/warranty_terms
•
•
Three years for the product
Three months for the product's standard accessories, unless otherwise
specified
Please take note that for the product, the warranty does not cover:
•
•
•
Damage from contamination
Normal wear and tear of mechanical components
Manuals, fuses, and standard disposable batteries
CALIBRATION CYCLE
One year
148
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Specifications
8
Measurement Category
The Agilent U1251A and U1252A Handheld Digital
Multimeter has a safety rating of CAT III 1000 V/ CAT IV,
600 V.
Measurement category definition
Measurement CAT I are measurements performed on circuits
which are not directly connected to the AC mains. For
example, measurements on circuits not derived from the AC
mains or specially protected (internal) mains-derived
circuits.
Measurement CAT II are measurements performed on circuits
that are directly connected to a low voltage installation. For
example, measurements on household appliances, portable
tools and other similar equipment.
Measurement CAT III are measurements performed on
building installations. For example, measurements on
distribution boards, circuit-breakers, wiring (including
cables), bus-bars, junction boxes, switches, socket outlets
within the fixed installation, equipment for industrial use
and equipment that is permanently connected to the fixed
installation such as stationary motors.
Measurement CAT IV are measurements performed at the
source of the low-voltage installation. For example,
measurements on electricity meters, primary over-current
protection devices and ripple control units.
Agilent U1251A/U1252A User’s and Service Guide
149
Download from Www.Somanuals.com. All Manuals Search And Download.
8
Specifications
Specification Assumptions
• The DC specifications are defined for measurements
which are taken after at least 1 minute of warm-up time.
• The AC and AC+DC specifications are defined for
measurements of sine wave and are taken after at least 1
minute of warm-up time.
• The accuracy of the multimeter may be affected when
making measurements in an environment where
electromagnetic interferences or significant electrostatic
charges are present.
150
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Specifications
8
Electrical Specifications
DC Specifications
Accuracy
Test Current/
Function
Range
Resolution
0.001 mV
0.01 mV
0.1 mV
0.0001 V
0.001 V
0.01 V
Burden Voltage
U1251A
0.05 + 50
U1252A
0.05 + 50
[2]
[2]
50.000 mV
500.00 mV
1000.0 mV
5.0000 V
50.000 V
500.00 V
1000.0 V
-
-
-
-
-
-
-
[1]
Voltage
0.025 + 5
0.03 + 5
0.03 + 5
0.1 V
Notes for DC voltage specifications:
1
2
Input impedance: >1GΩ for 50 mV to 1000 mV ranges. For U1251A, input impedance is 10 MΩ (nominal) for 5 V to 1000 V
The accuracy could be 0.05 % +10 for U1251A and 0.05 % + 5 for U1252A. Always use the Null function to zero out the
thermal effect before measuring the signal.
Accuracy
Test Current/
Function
Range
500.00 Ω
5.0000 kΩ
50.000 kΩ
500.00 kΩ
5.0000 MΩ
Resolution
0.01 Ω
0.0001 kΩ
0.01 kΩ
Burden Voltage
U1251A
0.08 + 10
U1252A
0.05 + 10
[3]
[3]
1.04 mA
416 µA
41.2 µA
4.12 µA
375 nA
187 nA
0.08 + 5
0.05 + 5
0.0001 MΩ
0.001 MΩ
0.2 + 5
1 + 10
0.15 + 5
1 + 5
Resistance
[4]
50.000 MΩ
3 + 10 < 200MΩ/
[4]
500.00 MΩ
0.01 MΩ
187 nA
187 nA
-
8 + 10 > 200MΩ
[5]
500.00 nS
Notes for resistance specifications:
0.01 nS
1 + 20
1 + 10
3
The accuracy of 500 Ω and 5 kΩ is specified after Null function, which is used to subtract the test lead resistance and the
thermal effect.
4
5
For the range of 50/500 MΩ, the R.H. is specified for < 60 %.
The accuracy is specified for < 50 nS and after Null function with open test lead.
Agilent U1251A/U1252A User’s and Service Guide
151
Download from Www.Somanuals.com. All Manuals Search And Download.
8
Specifications
Table 7-1 DC Accuracy (% of reading + Number of Least Significant Digit) (continued)
Accuracy
Test Current/
Function
Range
Resolution
0.01 µA
Burden Voltage
U1251A
U1252A
0.05 + 5
0.05 + 5
0.15 + 5
0.15 + 5
[6]
[6]
[6]
[6]
[6]
500.00 µA
5000.0 µA
50.000 mA
440.00 mA
5.0000 A
0.06 V
0.6 V
0.09 V
0.9 V
0.2 V
0.4 V
0.1 + 5
0.1 + 5
0.2 + 5
0.2 + 5
[6]
[6]
[6]
0.1 µA
0.001 mA
0.01 mA
0.0001 A
0.001 A
Current
0.3 + 10
0.3 + 10
0.3 + 10
0.3 + 5
[7]
10.000 A
Notes for current specifications:
6
Always use the Null function to zero out thermal effect with open test lead before measuring the signal. If the Null
function is not used, add 20 counts to the specified accuracy. Thermal effect could occur due to the following:
•
Wrong operation – where the resistance, diode or mV measurement function is used to measure high voltage signals
within the range of 50 V to 1000V
•
•
After battery-charging is complete.
After measuring a current greater than 440 mA. Thus we recommend that the meter be left to cool down for twice the
length of measurement time used.
7
Current can be measured up to 10 A continuously. An additional 0.5 % needs to be added to the specified accuracy if the
signal measured is in the range of 10 A to 20 A 30 seconds maximum. After measuring current > 10 A, leave the meter to
cool down (in switched OFF state) for twice the length of measuring time used before application of low current
measurement.
Accuracy
Test Current/
Function
Range
Resolution
Burden Voltage
U1251A
U1252A
[8]
Diode
-
0.1 mV
1.04 mA
0.05 + 5
Notes for diode specifications:
8
Maximum open voltage: < + 4.2 V.
152
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Specifications
8
AC Specifications
AC specifications for U1251A
Frequency
Function
Range
Resolution
0.001 mV
0.01 mV
0.1 mV
0.0001 V
0.001 V
0.01 V
30 Hz to 45 Hz
1 + 60
45 Hz to 1 kHz
0.6 + 40
0.6 + 25
0.6 + 25
0.6 + 25
0.6 + 25
0.6 + 25
0.6 + 40
1 kHz to 10 kHz
1.0 + 40
10 kHz to 30 kHz
1.6 + 60
50.000 mV
500.00 mV
1000.0 mV
5.0000 V
50.000 V
500.00 V
1000.0 V
1 + 60
1 + 60
1 + 60
1 + 60
1 + 60
1 + 60
1.0 + 40
1.0 + 25
1.0 + 25
1.0 + 25
1.0 + 25
1.0 + 40
1.6 + 60
1.6 + 40
1.6 + 40
[1]
Voltage
1.6 + 40
[2]
1.6 + 40
0.1 V
-
Notes for U1251A AC voltage specifications:
1
2
Input impedance: > 1GHz for 50 mV to 1000 mV. Input impedance is 1.1 MΩ (nominal) with <100 pF for 5 V to 1000 V
ranges.
The input signal is lower than the product of 20,000,000 V-Hz (product of voltage and frequency).
Table 7-3 U1251A accuracy specifications (% of reading + number of LSD) for true RMS AC current
Frequency
Function
Range
Resolution
0.01 µA
30 Hz to 45 Hz
1.5 + 50
45 Hz to 2 kHz
0.8 + 20
2 kHz to 20 kHz
3 + 80
[1]
500.00 µA
5000.0 µA
50.000 mA
440.00 mA
5.0000 A
0.1 µA
1.5 + 40
1.5 + 40
1.5 + 40
0.8 + 20
0.8 + 20
0.8 + 20
0.8 + 20
3 + 60
3 + 60
3 + 60
3 + 60
Current
0.001 mA
0.01 mA
0.0001 A
0.001 A
[3]
2 + 40
[2]
[3]
10.000 A
2 + 40
0.8 + 20
< 3 A/5 kHz
Notes for U1251A AC current specifications:
1
2
Input current > 35 µArms.
Current can be measured from 2.5 A up to 10 A continuously. Add 0.5% to the specified accuracy if the signal measured is
in the range of 10 A to 20 A and for a period of up to 30 seconds. After measuring current > 10 A, leave the meter to cool
down for a period that is twice the measuring time used before application of low current measurement.
Input current < 3 Arms.
3
Agilent U1251A/U1252A User’s and Service Guide
153
Download from Www.Somanuals.com. All Manuals Search And Download.
8
Specifications
AC specifications for U1252A
Table 7-4 U1252A accuracy specifications (% of reading + number of LSD) for true RMS AC voltage
Frequency
20 Hz –
45 Hz
45 Hz –
1 kHz
1 kHz –
10 kHz
10 kHz –
20 kHz
20 kHz –
100 kHz
[1]
Function
Range
Resolution
0.001 mV
0.01 mV
0.1 mV
0.0001 V
0.001 V
0.01 V
50.000 mV
500.00 mV
1000.0 mV
5.0000 V
50.000 V
500.00 V
1000.0 V
1.5 + 60
1.5 + 60
1.5 + 60
1.5 + 60
1.5 + 60
1.5 + 60
1.5 + 60
0.4 + 40
0.4 + 25
0.4 + 25
0.4 + 25
0.4 + 25
0.4 + 25
0.4 + 40
0.7 + 40
0.4 + 25
0.4 + 25
0.4 + 25
0.4 + 25
0.4 + 25
0.4 + 40
0.75 + 40
0.75 + 40
0.75 + 40
0.75 + 40
0.75 + 40
1.5 + 40
3.5 +120
3.5 + 120
3.5 + 120
3.5 + 120
[2]
Voltage
[3]
3.5 + 120
-
[3]
0.1 V
1.5 + 40
Notes for U1252A AC voltage specifications:
1
2
The additional error to be added as frequency > 20 kHz and signal input <10 %of range: 3 counts of LSD per kHz.
Input impedance: > 1GHz for 50 mV to 1000 mV. Input impedance is 1.1 MΩ (nominal) with <100 pF for 5 V to 1000 V
ranges.
3
The input signal is lower than the product of 20,000,000 V-Hz (product of voltage and frequency).
Table 7-5 U1252A accuracy specifications (% of reading + number of LSD) for true RMS AC current
Frequency
20 kHz –
100 kHz
20 Hz – 45 Hz 45 Hz – 1 kHz 1 kHz – 20 kHz
[1]
Function
Range
Resolution
0.01 µA
[2]
500.00 µA
1.0 + 20
1.0 + 20
1.0 + 20
1.0 + 20
0.7 + 20
0.7 + 20
0.7 + 20
0.7 + 20
0.7 + 20
0.7 + 20
0.75 + 20
0.75 + 20
0.75 + 20
1.5 + 20
3 + 60
5 + 80
5 + 80
5 + 80
5 + 80
5000.0 µA
50.000 mA
440.00 mA
5.0000 A
0.1 µA
Current
0.001 mA
0.01 mA
0.0001 A
0.001 A
[4]
1.5 + 20
1.5 + 20
-
[3]
[4]
10.000 A
< 3 A/5 kHz
Notes for U1252B AC current specifications:
1
2
3
The additional error to be added as frequency > 20 kHz and signal input< 10 % of range: 3 counts of LSD per kHz.
Input current > 35 µArms.
Current can be measured from 2.5 A up to 10 A continuously. Add 0.5% to the specified accuracy if the signal measured is
in the range of 10 A to 20 A and for a period of up to 30 seconds. After measuring current > 10 A, leave the meter to cool
down for a period that is twice the measuring time used before application of low current measurement.
Input current < 3 Arms.
4
154
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Specifications
8
AC+DC Specifications for U1252A
Table 7-6 U1252A true RMS ac+dc voltage specifications
Frequency
1 kHz –
10 kHz
0.7 + 60
0.4 + 30
0.4 + 30
0.4 + 30
0.4 + 30
0.4 + 30
0.4 + 45
30 Hz –
45 Hz
45 Hz –
1 kHz
10 kHz –
20 kHz
20kHz –
100kHz
[1]
Function
Range
Resolution
0.001 mV
0.01 mV
0.1 mV
0.0001 V
0.001 V
0.01 V
50.000 mV
500.00 mV
1000.0 mV
5.0000 V
50.000 V
500.00 V
1000.0 V
1.5 + 80
1.5 + 65
1.5 + 65
1.5 + 65
1.5 + 65
1.5 + 65
1.5 + 65
0.4 + 60
0.4 + 30
0.4 + 30
0.4 + 30
0.4 + 30
0.4 + 30
0.4 + 45
0.8 + 60
0.8 + 45
0.8 + 45
0.8 + 45
0.8 + 45
1.5 + 45
3.5 + 125
3.5 + 125
3.5 + 125
3.5 + 125
[2]
Voltage
[3]
3.5 + 125
-
[3]
0.1 V
1.5 + 45
Notes for U1252A AC +DC voltage specifications:
1
2
The additional error to be added as frequency > 20 kHz and signal input <10 %of range: 3 counts of LSD per kHz.
Input impedance: > 1 GW for 50 mV to 1000 mV. Input impedance is 1.1 MΩ (nominal) with <100 pF for 5 V to 1000 V
ranges.
3
The input signal is lower than the product of 20,000,000 V-Hz (product of voltage and frequency).
Table 7-7 U1252A true RMS ac+dc current specifications
Frequency
45 Hz – 1 kHz
0.8 + 25
Function
Range
Resolution
0.01 µA
30 Hz – 45 Hz
1.1 + 25
1 kHz – 20 kHz
0.8 + 25
[1]
500.00 µA
5000.0 µA
50.000 mA
440.00 mA
5.0000 A
0.1 µA
1.1 + 25
1.2 + 25
1.2 + 25
0.8 + 25
0.9 + 25
0.9 + 25
0.9 + 30
0.8 + 25
0.9 + 25
0.9 + 25
3.3 + 70
0.001 mA
0.01 mA
0.0001 A
0.001 A
Current
[3]
1.8 + 30
[2]
[3]
10.000 A
1.8 + 30
0.9 + 25
< 3 A/5 kHz
Notes for U1252A AC +DC current specifications:
1
2
Input current > 35 µArms.
Current can be measured from 2.5 A up to 10 A continuously. Add 0.5% to the specified accuracy if the signal measured is
in the range of 10 A to 20 A and for a period of up to 30 seconds. After measuring current > 10 A, leave the meter to cool
down for a period that is twice the measuring time used before application of low current measurement.
Input current < 3 Arms.
3
Agilent U1251A/U1252A User’s and Service Guide
155
Download from Www.Somanuals.com. All Manuals Search And Download.
8
Specifications
Capacitance Specifications
Table 7-8 Capacitance specifications
Accuracy
(% of reading +
Offset Error)
1 % + 8
Display update rate
(approx)
Function
Range
Resolution
10.000 nF
100.00 nF
1000.0 nF
10.000 µF
100.00 µF
1000.0 µF
10.000 mF
100.00 mF
0.001 nF
0.01 nF
0.1 nF
0.001 µF
0.01 µF
0.1 µF
4 times/sec.
Capacitance
1 % + 5
1 time/sec.
0.1 times/sec.
0.01 times/sec.
0.001 mF
0.01 mF
3 % + 10
Temperature Specifications
Table 7-9 Temperature specifications
Accuracy
Thermocouple
Range
Resolution
(% of reading + No. of
Least Significant Digit)
0.3 % + 3 °C/
Type
–200 – 1372 °C/
0.1 °C/
K
–328 – 2502 °F
0.1 °F
0.3 % + 6 °F
–210 – 1200 °C/
0.1 °C/
0.3 % + 3 °C/
[2]
J
–346 – 2192 °F
0.1 °F
0.3 % + 6 °F
Notes for temperature specifications:
1
Accuracy of specifications is subject to the following conditions:
•
The accuracy does not include the tolerance of the thermocouple probe. The thermal sensor plugged into the meter
should be placed in the operating environment for at least an hour.
•
Use the Null function to reduce the thermal effect. Before using the NULL function, set the meter to no ambient
compensation (
) mode and keep the thermocouple probe as close to the meter as possible, avoiding contact with
any surface that has a different temperature from the ambient temperature.
•
When measuring temperature with respect to any temperature calibrator, try to set both the calibrator and the meter
with an external reference (without internal ambient compensation). If both the calibrator and the meter are set with
internal reference (with internal ambient compensation), deviation may show between the readings of the calibrator
and the meter, due to differences in ambient compensation between the calibrator and the meter.
2
This thermocouple type is only available for the U1252A.
156
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Specifications
8
Frequency Specifications
Table 7-10 Frequency specifications
Accuracy
Range
Resolution
(% of reading + No. of
Least Significant Digit)
Min. Input Frequency
99.999 Hz
999.99 Hz
9.9999 kHz
99.999 kHz
999.99 kHz
0.001 Hz
0.01 Hz
0.0001 kHz
0.001 kHz
0.01 kHz
0.02% + 3
< 600 kHz
1 Hz
Notes for frequency specifications:
1
The multimeter will automatically select the most appropriate range when making frequency measurements.
Duty Cycle and Pulse Width Specifications
Table 7-11 Duty cycle and pulse width specifications
Function
Mode
Range
0.01% to 99.99%
5% to 95%
Resolution
Accuracy (at full scale)
DC Coupling
AC Coupling
-
-
0.3% per kHz + 0.3%
0.3% per kHz + 0.3%
Duty cycle
Notes for duty cycle specifications:
1
2
The accuracy for duty cycle and pulse width is based on a 5 V square wave input into the DC 5 V range.
For AC coupling, the duty cycle range can be measured for signal frequency > 20 Hz.
-
-
500 ms
0.01 ms
0.1 ms
0.2% + 3
0.2% + 3
Pulse width
2000 ms
Notes for pulse width specifications:
1
2
The accuracy for duty cycle and pulse width is based on a 5 V square wave input into the DC 5 V range.
Positive or negative pulse width must be greater than 10 µs and the range of duty cycle should be considered. The range
of pulse width is determined by the frequency of the signal.
Agilent U1251A/U1252A User’s and Service Guide
157
Download from Www.Somanuals.com. All Manuals Search And Download.
8
Specifications
Frequency Sensitivity Specifications
For voltage measurements
Table 7-12 Frequency sensitivity and trigger level specifications for voltage measurements
Minimum sensitivity (RMS sine wave)
Trigger level for DC coupling
Model Number
Input
range
[1]
U1251A
U1252A
U1251A
U1252A
20 Hz - 100
kHz
>100 kHz - 20 Hz - 200 >200 kHz -
>100 kHz -
200 kHz
> 100 kHz -
500 kHz
< 100 kHz
< 100 kHz
200 kHz
15 mV
35 mV
50 mV
kHz
10 mV
70 mV
120 mV
500 kHz
25 mV
50.000 mV
500.00 mV
1000.0 mV
10 mV
25 mV
40 mV
10 mV
60 mV
100 mV
15 mV
10 mV
70 mV
120 mV
25 mV
150 mV
300 mV
70 mV
150 mV
300 mV
150 mV
0.5 V/1.25
V (< 100
Hz)
5.0000 V
0.25 V
0.5 V
0.3 V
1.2 V
0.6 V
0.6 V
1.5 V
50.000 V
500.00 V
2.5 V
25 V
5 V
-
3 V
5 V
-
5 V
6 V
-
6 V
15 V
-
30 V < 100
kHz
50 V
60 V
50 V < 100
kHz
1000.0 V
50 V
-
-
300 V
-
120 V
-
Notes for frequency sensitivity and trigger level specifications for voltage measurements;
1
2
Maximum input for specified accuracy = 10 x range or 1000 V.
The input signal is lower than the product of 20,000,000 V-Hz.
158
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Specifications
8
For current measurements
Table 7-13 Frequency sensitivity specifications for current measurements
Minimum Sensitivity (R.M.S. sine wave)
Input Range
20 Hz – 20 kHz
100 µA
250 µA
10 mA
500.00 µA
5000.0 µA
50.000 mA
440.00 mA
5.0000 A
25 mA
1 A
Peak Hold Specifications
Table 7-14 Peak hold specifications for dc voltage and current measurements
Signal Width
Accuracy For DC mV/voltage/current
2% + 400 for all ranges
Single event > 1 ms
Repetitive > 250 µs
2% + 1000 for all ranges
Agilent U1251A/U1252A User’s and Service Guide
159
Download from Www.Somanuals.com. All Manuals Search And Download.
8
Specifications
Frequency Counter Specifications for U1252A
Table 7-15 Frequency counter (divide by 1) specifications
Accuracy
(% of reading + No.
Range
Resolution
Sensitivity
Min. Input Frequency
of Least Significant
Digit)
[1]
99.999 Hz
999.99 Hz
9.9999 kHz
99.999 kHz
999.99 kHz
9.9999 MHz
0.001 Hz
0.01 Hz
0.0001 kHz
0.001 kHz
0.01 kHz
0.02% + 3
100 mV R.M.S.
200 mV R.M.S.
0.5 Hz
0.002% + 5
< 985 kHz
0.0001 MHz
[4]
Table 7-16 Frequency counter (divide by 100 ) specifications
Accuracy
(% of reading + No.
of Least Significant
Digit)
Range
Resolution
Sensitivity
Min. Input Frequency
9.9999 MHz
99.99 MHz
0.0001 MHz
0.001 MHz
0.002 % + 5,
400 mV R.M.S.
600 mV R.M.S.
1 MHz
< 20 MHz
Notes for frequency counter specifications:
1
All frequency counters are susceptible to error when measuring low-voltage, low-frequency signals. Shielding inputs from
picking up external noise is critical for minimizing measurement errors. For non-square wave signals, an additional 5
counts need to be added.
2
3
4
The maximum measurement level is < 30 Vpp.
The minimum measurement frequency of low frequency is set by power-on option to speed up the measurement rate.
Shown on the secondary display
160
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Specifications
8
Square Wave Output for U1252A
[1]
Output
Range
Accuracy
0.5, 1, 2, 5, 10, 15, 20, 25, 30, 40, 50, 60,
75, 80, 100, 120, 150, 200, 240, 300,
400, 480, 600, 800, 1200, 1600, 2400,
4800 Hz
Frequency
0.005% x output frequency + 2 counts
[2][4][5]
[3]
Duty Cycle
0.39% – 99.60%
1/Frequency
0.398% of full scale
[2][4][6]
Pulse Width
0.2 ms + (range/256)
0.2 V
Amplitude
Notes for square wave output specifications:
Fixed 0 to + 2.8 V
1
2
Output impedance: 3.5 kW maximum.
The positive or negative pulse width must be greater than 50 µs for adjusting the duty cycle or the pulse width under
different frequencies. Otherwise the accuracy and the range will be different from the definition.
For signal frequencies greater than 1 kHz, add 0.1% per kHz to the accuracy.
The accuracy for duty cycle and pulse width is based on a 5 V square wave input without dividing signal.
Duty cycle can be set for 256 steps and each step is 0.390625% per kHz.
3
4
5
6
The pulse width can be set for 256 steps and each step is 1/(256 x Frequency).
Agilent U1251A/U1252A User’s and Service Guide
161
Download from Www.Somanuals.com. All Manuals Search And Download.
8
Specifications
Operating Specifications
Display update rate (approximate)
Table 7-18 Display update rate (approximate)
Function
AC V
Times/second
7
AC V + dB
7
DC V (V or mV)
AC V (V or mV)
AC+DC V (V or mV)
Ω / nS
7
7
2
14
Diode
14
Capacitance
4 (< 100 µF)
DC A ( µA, mA, or A)
AC A ( µA, mA, or A)
AC+DC A ( µA, mA, or A)
Temperature
Frequency
7
7
2
6
2 (> 10 Hz)
1 (> 10 Hz)
1 (> 10 Hz)
Duty cycle
Pulse width
The U1251A and U1252A handheld digital multimeter does not contain a
realtime clock. Only ONE sample per second can be logged.
NOTE
162
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Specifications
8
Input impedance
Table 7-19 Input Impedance
Function
Range
Input Impedance
50.000 mV
500.00 mV
1000.0 mV
5.0000 V
50.000 V
500.00 V
1000.0 V
50.000 mV
500.00 mV
1000.0 mV
5.0000 V
50.000 V
500.00 V
1000.0 V
50.000 mV
500.00 mV
1000.0 mV
5.0000 V
50.000 V
500.00 V
1000.0 V
10.00 MΩ
10.00 MΩ
10.00 MΩ
[1]
DC Voltage
11.10 MΩ
10.10 MΩ
10.01 MΩ
10.001 MΩ
10.00 MΩ
10.00 MΩ
10.00 MΩ
[2]
AC Voltage
10.00 MΩ
10.00 MΩ
10.00 MΩ
10.00 MΩ
10.00 MΩ
10.00 MΩ
10.00 MΩ
[2]
AC + DC Voltage
11.10 MΩ // 10 MΩ
10.10 MΩ // 10 MΩ
10.01 MΩ // 10 MΩ
10.001 MΩ // 10 MΩ
Notes for input impedance:
1
2
For 5 V to 1000 V range, the specified input impedance in parallel with 10 MΩ at dual display.
The specified input impedance (nominal) in parallel with <100 pF.
Agilent U1251A/U1252A User’s and Service Guide
163
Download from Www.Somanuals.com. All Manuals Search And Download.
8
Specifications
164
Agilent U1251A/U1252A User’s and Service Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
www.agilent.com
Contact us
To obtain service, warranty or technical support
assistance, contact us at the
following phone numbers:
United States:
(tel) 800 829 4444
Canada:
(tel) 877 894 4414
China:
(tel) 800 810 0189
Europe:
(fax) 800 829 4433
(fax) 800 746 4866
(fax) 800 820 2816
(tel) 31 20 547 2111
Japan:
(tel) (81) 426 56 7832 (fax) (81) 426 56 7840
Korea:
(tel) (080) 769 0800
Latin America:
(tel) (305) 269 7500
Taiwan:
(fax) (080) 769 0900
(tel) 0800 047 866
(fax) 0800 286 331
Other Asia Pacific Countries:
(tel) (65) 6375 8100
(fax) (65) 6755 0042
Or visit Agilent worlwide Web at:
Product specifications and descriptions in this
document are subject to change without notice.
Always refer to the Agilent Web site for the latest
revision.
© Agilent Technologies, Inc. 2006 – 2012
Printed in Malaysia
Tenth Edition, May 4, 2012
U1251-90003
Agilent Technologies
Download from Www.Somanuals.com. All Manuals Search And Download.
|
Advantech Computer Hardware UNO 2173A AF User Manual
AEG Cooktop FM 6300G AN User Manual
Agilent Technologies Video Games 66319b User Manual
Aiwa Stereo System NSX DS8 User Manual
AKG Acoustics Power Supply B 29L User Manual
Allied Telesis Webcam management software layer 2+ fast ethernet switches User Manual
Amana Clothes Dryer NGD7300W User Manual
American Standard Outdoor Shower 6032Y1SW User Manual
Bakers Pride Oven Outdoor Storage CA71 IN39 User Manual
Beko Hair Dryer BKK 2168 User Manual