4
Faults
Monitor operation
The air-tightness of the air-pouch
system is checked regularly. If a leak
is detected, the red warning
indicator shows alarm. When the
leak is removed the red warning
indicator disappears and the
After placement of probe the tube
connector is inserted into the
monitor socket and fastened by a
slight clockwise turn. The soft
material of the connector
guarantees airtightness.
measurement is continued. If the
leak has not been removed after
one minute, in addition to the alarm
display a chirp sound is emitted. To
reset the alarm the monitor must be
switched off and on again. The
chirp sound stops after one minute,
unless the monitor has been
switched off before, and the digital
display dissapears. The warning
indicator continues to flash.
Pressure measurements are started
by switching on the monitor. No
manual zero adjustment or
calibration is required.
If a pressure limit is exceeded during
probe filling the filling process is
interrupted, the digital display
shows ´E1´ and the chirp sound is
emitted. This can happen if the air-
pouch resists unfolding or if the
tube is clamped. Switching off and
on the monitor resets the system.
If the digital display shows ´E2´ or
´E3´ during operation an internal
fault has been detected. Contact
the service in this case.
Warning
The unit cannot be
repaired by the user.
5
6
Connections
On the digital display the mean
pressure is shown in mmHg.
Connection of a Patient
Monitor
Connection of a Strip-Chart
Recorder
To connect a patient monitor the
connecting cable is inserted into the
socket of the IAP-Monitor and
connected to the pressure
transducer input of the patient
monitor. Connect probe to the IAP-
Monitor. Switch on the Brain-
Pressure Monitor. Wait for the
digital display of the mean ICP. The
pulsatile signal is displayed now on
the monitor screen.
To connect a strip-chart recorder
with voltage input the appropriate
cable is connected with the socket
at the rear panel of the monitor and
the input socket of the recorder.
During the first ten minutes after
switching on the pulse amplitude
(i.e. systolic pressure minus diastolic
pressure) is additionally displayed
marked by an ´A´.
Connection of a Computer
The RS 232 C interface permits
communication with any computer.
During this time the mean pressure
is displayed for three seconds, then
the amplitude is displayed for one
second.
Zero Adjustment of the
Patient Monitor
Disconnect the catheter from the
IAP-Monitor. Leave the IAP-Monitor
and the patient monitor switched
on. Zero patient monitor using the
zero button. Reconnect the probe.
Switch off the IAP-Monitor. Switch
it on again. Observe the display of
the mean ICP on the IAP-Monitor
and the pulsatile signal on the
patient monitor.
Cleaning
After removal of the power cord the
unit can be wiped off with a damp
rag using a common disinfectant.
No liquid whatsoever must be
Warning
Unplug the power cord
prior to cleaning.
Connection of a CPP-Monitor
Connect the IAP-Monitor to the RS
232 IAP socket of the CPP-Monitor.
introduced into the catheter or into
the connector of the monitor.
Warning
No fluid should be
allowed to enter the
probe connector.
Warning
In case of unintentional
wettening of the device,
the power cord is to be
disconnected immediately.
3
8
Indication and Method
Technical specifications
The IAP-Monitor and the catheters
are intended for the monitoring of
intraabdominal pressure in the
stomach.
REF
IAP 32.1
Measurement range
Operating voltage
-50 to +100 mmHg
230 V~, 50/60 Hz
or
115 V~, 50/60 Hz
The air-pouch system consists of a
hollow body connected to a
pressure transducer by tubing. The
pressure transducer, the electronic
hardware, and the device for filling
the air-pouch are integrated in
the IAP-Monitor.
Current requirement
Fuses
0,1 A at 230 V~
0,2 A at115 V~
125 mA T at 230 V~
250 mA T at 115 V~
For intraabdominal pressure
measurements the air- pouch is
placed in the stomach of the
Voltage output
1V/50 mmHg at 1 kΩ
patient. The intraabdominal pressure
is transmitted across the thin pouch
wall to the air volume in the pouch
and transformed into an electric
signal by the pressure transducer.
Pressure monitor output
RS 232 C interface
5 µV/mmHg/V
9600 Bd (other baud rates
upon request)
Accuracy class
2
On the digital display the mean
pressure is shown. At the monitor
output both the mean pressure and
the pulsatile signal are available.
Safety class
I / BF
3 kg
Weight
Once every hour the IAP-Monitor
opens the pressure transducer to
atmospheric pressure for zero
adjustment. The air-pouch is
then filled with the exact air volume
required for accurate pressure
transmission.
In accordance with standards
IEC 60601-1
IEC 60601-1-2
1
10
Accessories
1.9.2003 AS/az
Description
CPP-Monitor
IAP-Catheter
Order number
CPP 21.x
EG-Konformitätserklärung
Declaration of Conformity
IAP 32.1.11
KBL 21.005.00
Das Produkt
Interconnect Cable for
IAP-Monitor HDM IAP 32.1
Spiegelberg CPP-Monitor 21.x
entspricht den grundlegenden Anforderungen gemäß Anhang
I der Richtlinie 93/42/EWG. Es ist als Klasse IIa klas-
sifiziert.
Monitor Cables
Bruker VM800/900
Corometrix
Datex-Cardiocap
Dräger VM 300
Hellige 4. Generation up
Hellige 3. Generation
Hewlett-Packard
Horizon
KBL 13.014.00
KBL 13.020.00
KBL 13.007.00
KBL 13.001.00
KBL 13.003.00
KBL 13.019.00
KBL 13.004.00
KBL 13.015.00
KBL 13.018.00
KBL 13.005.00
KBL 13.005.01
KBL 13.008.00
KBL 13.008.01
KBL 13.011.00
KBL 13.009.00
KBL 13.010.00
KBL 13.002.00
KBL 13.006.00
KBL 13.013.00
Das Qualitätsmanagementsystem entspricht den Forderungen
der Norm ISO 13485:2001 gemäß Zertifikat Nr. 18737 der
DQS, benannte Stelle Nr. 0297.
The Product
IAP-Monitor IAP 32.1
is in accordance with the requirements of Annex I of the
Manual Devices Directive 93/42/EEC. It is classified as
class IIa.
Kontron
Marquette
Marquette 7010
Nihon-Kohden
Nihon-Kohden BMS 8300
Philips CM130
Propaque/Mennen
Schiller CM-8
Siemens-Sirecust
Space-Labs
S&W, Danica
The Quality Management System is in accordance with the
requirements of the standard ISO 13485:2001 according to
certificate no. 18737 of DQS, Notified Body No. 0297.
Spiegelberg (GmbH & Co.) KG
Recorder Cables for Recorders
with Voltage Input
BNC-Plug
KBL 13.030.00
KBL 13.031.00
Dr. A. Spiegelberg
Banana Plug 4 mm
Computer Cable RS 232 9-pin
KBL 13.033.00
|