Agilent 83491/2/3A
Clock Recovery Modules
User’s Guide
General Safety Considerations
General Safety Considerations
This product has been designed and tested in accordance with IEC Publica-
tion 61010-1, Safety Requirements for Electrical Equipment for Measurement,
Control and Laboratory Use, and has been supplied in a safe condition. The
instruction documentation contains information and warnings that must be
followed by the user to ensure safe operation and to maintain the product in a
safe condition.
W A R N I N G
Light energy can radiate from the front panel OUTPUT connectors on
Agilent 83492A and 83493A modules. The light emitted from these
connectors is the slightly attenuated light that is input to the front-
panel INPUT connector.
W A R N I N G
W A R N I N G
If this instrument is not used as specified, the protection provided by
the equipment could be impaired. This instrument must be used in a
normal condition (in which all means for protection are intact) only.
To prevent electrical shock, disconnect the Agilent 83491/2/3A from
mains before cleaning. Use a dry cloth or one slightly dampened with
water to clean the external case parts. Do not attempt to clean
internally.
W A R N I N G
This is a Safety Class 1 product (provided with a protective earthing
ground incorporated in the power cord). The mains plug shall only be
inserted in a socket outlet provided with a protective earth contact.
Any interruption of the protective conductor inside or outside of the
product is likely to make the product dangerous. Intentional
interruption is prohibited.
W A R N I N G
W A R N I N G
No operator serviceable parts inside. Refer servicing to qualified
personnel. To prevent electrical shock, do not remove covers.
For continued protection against fire hazard, replace line fuse only
with same type and ratings, (type T 0.315A/250V for 100/120V
operation and 0.16A/250V for 220/240V operation). The use of other
fuses or materials is prohibited. Verify that the value of the line-
voltage fuse is correct.
• For 100/120V operation, use an IEC 127 5×20 mm, 0.315 A, 250 V, Agilent
iii
General Safety Considerations
Technologies part number 2110-0449.
• For 220/240V operation, use an IEC 127 5×20 mm, 0.16 A, 250 V, Agilent
Technologies part number 2110-0448.
C A U T I O N
Before switching on this instrument, make sure that the line voltage selector
switch is set to the line voltage of the power supply and the correct fuse is
installed. Assure the supply voltage is in the specified range.
C A U T I O N
C A U T I O N
This product is designed for use in Installation Category II and Pollution
Degree 2 per IEC 1010 and 664 respectively.
VENTILATION REQUIREMENTS: When installing the product in a cabinet, the
convection into and out of the product must not be restricted. The ambient
temperature (outside the cabinet) must be less than the maximum operating
temperature of the product by 4°C for every 100 watts dissipated in the
cabinet. If the total power dissipated in the cabinet is greater than 800 watts,
then forced convection must be used.
C A U T I O N
Always use the three-prong ac power cord supplied with this instrument.
Failure to ensure adequate earth grounding by not using this cord may cause
instrument damage.
C A U T I O N
C A U T I O N
C A U T I O N
Do not connect ac power until you have verified the line voltage is correct.
Damage to the equipment could result.
This instrument has autoranging line voltage input. Be sure the supply voltage
is within the specified range.
Electrostatic discharge (ESD) on or near input connectors can damage circuits
inside the instrument. Repair of damage due to misuse is not covered under
warranty. Before connecting any cable to the electrical input, momentarily
short the center and outer conductors of the cable together. Personnel should
be properly grounded, and should touch the frame of the instrument before
touching any connector.
iv
Contents
1 Installation
Installation 1-2
2 Operation
Agilent 83491/2/3A Modules—At a Glance 2-2
Block Diagrams 2-7
3 Specifications and Regulatory Information
4 Reference
In Case of Difficulty 4-3
Error Messages 4-5
Electrostatic Discharge Information 4-8
Cleaning Connections for Accurate Measurements 4-10
Returning the Instrument for Service 4-20
Agilent Technologies Service Offices 4-23
Contents-1
Installation
Installation
Installation
Agilent 83491/2/3A modules require that firmware revision A.06.25 or later be
installed in the Agilent 83480A. If you wish to install the module in an
Agilent 54750A digitizing oscilloscope, you must first install the
Agilent 83480K communications firmware upgrade kit.
To check the Agilent 83480A’s firmware revision code
1 Press the Utility key and then the System config softkey.
2 The firmware revision number is listed under the Frame section of the display.
C A U T I O N
Fiber-optic connectors are easily damaged when connected to dirty or
damaged cables and accessories. The Agilent 83492A and 83493A front-panel
input connectors are no exception. When you use improper cleaning and
handling techniques, you risk expensive instrument repairs, damaged cables,
and compromised measurements. Before you connect any fiber-optic cable to
an Agilent 83492A or 83493A module, refer to “Cleaning Connections for
Accurate Measurements” on page 4-10.
C A U T I O N
The circuits on electrical inputs and outputs can be damaged by electrostatic
discharge (ESD). Therefore, avoid applying static discharges to any front or
rear-panel electrical connector. Before connecting any coaxial cable to a front-
panel connector, momentarily short the center and outer conductors of the
cable together. Avoid touching the front-panel connectors without first
touching the frame of the instrument. Be sure that the instrument is properly
earth-grounded to prevent buildup of static charge. Refer to “Electrostatic
Discharge Information” on page 4-8.
1-2
Installation
Installation
To install the module
1 Verify that all system components ordered have arrived by comparing the
shipping forms to the original purchase order. Inspect all shipping containers.
The shipment includes:
❒ An Agilent 83491/2/3A Clock Recovery Module
❒ Fiber-optic adapter cable (Agilent 83492/3A module only)
❒ Two RF adapter cables (Agilent 83491A module only)
❒ Agilent 83491/2/3A User’s Guide (this book)
If your shipment is damaged or incomplete, save the packing materials and
notify both the shipping carrier and the nearest Agilent Technologies service
office. Agilent Technologies will arrange for repair or replacement of damaged
or incomplete shipments without waiting for a settlement from the transporta-
tion company. Notify the Agilent Technologies customer engineer of any prob-
lems.
1 Make sure that the serial number listed on the module’s rear-panel label
matches the serial number listed on the shipping document.
Figure 1-1. Serial number label
2 Install the Agilent 83491/2/3A module into the Agilent 83480A mainframe’s left
to ensure that the module is securely seated in the mainframe.
Note
Installing the module into the left slot ensures that the supplied adapter cable will fit.
See Figure 1-2.
1-3
Installation
Installation
Figure 1-2. Position of modules in the mainframe
3 Clean all optical interfaces as described in “Cleaning Connections for Accurate
Measurements” on page 4-10, before making measurements.
4 Perform the following steps if you’re installing an Agilent 83492/3A module:
Agilent 83492/3A module’s front-panel Input connector.
On Agilent 83492A module’s, the input connector used depends on the
wavelength of the input signal. Refer to “To connect cables to an
Agilent 83492A” on page 1-7.
5 Connect the supplied adapter cable as shown in Figure 1-3.
W A R N I N G
Light energy can radiate from the front panel OUTPUT connectors on
Agilent 83492A and 83493A modules. The light emitted from these
connectors is the slightly attenuated light that is input to the front-
panel INPUT connector.
6 Turn on the Agilent 83480A, and connect a modulated signal to the
Agilent 83491/2/3A module’s Input connector.
1-4
Installation
Installation
Figure 1-3. The adapter cable
C A U T I O N
Agilent 83491A Modules: Maximum safe signal input level is 5V. The input
circuits can also be damaged by electrostatic discharge (ESD). Before
connecting any coaxial cable to the connectors, momentarily short the center
and outer conductors of the cable together. Avoid touching the front-panel
input connectors without first touching the frame of the instrument. Be sure
that the instrument is properly earth-grounded to prevent buildup of static
charge.
C A U T I O N
7 On the Agilent 83480A, press the Trigger key and then the Source softkey. Then,
select trigger 2 (the Agilent 83491/2/3A module) for the trigger source.
8 On the Agilent 83491/2/3A module, repeatedly press the SELECT key until a
front-panel light indicates the data rate of the signal. See Figure 1-4 on
page 1-6.
Green and red data-rate lights
The data-rate indicator lights change color between red and green to show which data
rate is selected. A red light does not indicate a problem. A red light shows that the adja-
cent red data rate label is selected. A green light shows that the adjacent green data
rate label is selected. Repeatedly pressing the SELECT key cycles through the selections
in one color before switching to the opposite color. On Agilent 83491A modules for
example, the first selection cycle shows 155 Mb/s selected. The second section cycle
shows 1062 Mb/s selected.
1-5
Installation
Installation
To connect cables to an Agilent 83492A
On Agilent 83492A modules, the front-panel fiber-optic connectors reverse
input and output roles depending on the wavelength of the signal. Signals in
the 750 nm to 860 nm wavelength range are input to the left connector and
output from the right connector. Signals in the 1000 nm to 1600 nm wave-
length range are input to the right connector and output from the left connec-
tor.
Figure 1-5. Input and output connections versus wavelength
1-7
2
Block Diagrams 2-7
To Display a Signal 2-8
To Compensate for Module Insertion Loss 2-9
Using Probes with an Agilent 83491A 2-10
To compensate for a passive probe 2-11
To compensate an Agilent 54701A active probe 2-11
To compensate for other devices 2-12
Operation
Operation
Agilent 83491/2/3A Modules—At a Glance
Agilent 83491/2/3A Modules—At a Glance
The Agilent 83491/2/3A are designed to operate in an Agilent 83480A digital
communications analyzer. These modules recover clock and data information
at standard telecom and datacom rates. The resulting trigger signal is made
available to the Agilent 83480A mainframe via a connector located on the
module’s rear-panel. An external front-panel cable passes the data signal, with
some insertion loss, to the receiver module.
Table 2-1. Module Features
Agilent
Input Connector
Selectable Rates (Mb/s)
Module
83491A
83492A
83493A
155, 622, 1060, 1250, 2120, 2488, 2500
155, 622, 1060, 1250, 2120, 2488, 2500
155, 622, 1250, 2488, 2500
50Ω electrical
Multimode fiber (62.5/125 µm)
Single-mode fiber (9/125 µm)
Figure 2-1. An Agilent 83493A installed in an Agilent 83480A mainframe
2-2
Operation
Agilent 83491/2/3A Modules—At a Glance
W A R N I N G
Light energy can radiate from the front panel OUTPUT connectors on
Agilent 83492A and 83493A modules. The light emitted from these
connectors is the slightly attenuated light that is input to the front-
panel INPUT connector.
Use with Agilent 71603B or 71612B Error Performance Analyzers
The front-panel Data and Clock outputs provide electrical recovered clock and
regenerated data signals for simultaneous testing with other instruments,
such as the Agilent 71603B or 71612B error performance analyzers.
Multimode module and single-mode reference receivers
Agilent Technologies does not recommend using the Agilent 83492A multi-
mode module with single-mode reference receivers such as the
Agilent 83481A, 83482A, or 83485A,B modules. Connecting multimode to sin-
gle-mode fibers causes large reflections and insertion loss because of the
reduction of the optical fiber’s core from 62.5 µm to 9 µm.
Single-mode module and multimode reference receivers
It is acceptable to use an Agilent 83493A single-mode module with a multi-
mode reference receiver such as the Agilent 83486A module. This is true pro-
vided that single-mode fiber is connected to the Agilent 83493A module’s
front-panel INPUT connector.
What you won’t find on these modules....
Unlike other modules designed to be used with the Agilent 83480A digital communica-
tions analyzer, the Agilent 83491/2/3A modules do not include Channel keys or menus.
Also, there are no GPIB programming commands for these modules.
2-3
Operation
Front-Panel Features
Front-Panel Features
Figure 2-2. Agilent 83491/2/3A front panels
SELECT key
Pressing this key changes the modulation rate of the input signal. The recov-
ered and retimed clock trigger is sent to the mainframe. The Trigger On Data
selection is a bypass mode where the data stream directly triggers the main-
frame. Refer to “Block Diagrams” on page 2-7 to view a schematic of the nor-
mal and bypass paths.
2-4
Operation
Front-Panel Features
Green and red data-rate lights
The data-rate indicator lights change color between red and green to show which data
rate is selected. A red light does not indicate a problem. A red light shows that the adja-
cent red data rate label is selected. A green light shows that the adjacent green data
rate label is selected. Repeatedly pressing the SELECT key cycles through the selections
in one color before switching to the opposite color. On Agilent 83491A modules for
example, the first selection cycle shows 155 Mb/s selected. The second section cycle
shows 1062 Mb/s selected.
UNLOCKED
indicator
This light shows when clock recovery cannot be established on the signal. If a
clock rate is selected, the trigger output to the mainframe is disabled to pre-
vent free-run triggering. However in bypass mode (Trigger On Data selected),
triggering is not disabled. When the UNLOCKED light is on, you can establish a
trigger on the data input to the reference receiver.
Auxiliary outputs DATA connector: This connector provides a fully regenerated version of the
input signal. It is intended for monitoring purposes only and not for rigorous
eye mask compliance testing. The frequency response does not conform to the
requirements for eye mask testing as described in ITU-T G.957 and Bellcore
GR-253-CORE. On Agilent 83492A and 83493A modules, this port is ampli-
tude stabilized for input signals greater than approximately –23 dBm.
CLOCK connector: This connector provides the recovered clock signal. You can
use this signal to measure jitter transfer, because this output can track and fol-
low input data with very fast jitter; it has a wide bandwidth jitter transfer func-
tion when compared to the recovered clock signal which is routed through a
rear-panel connector to the mainframe for triggering. Note that the CLOCK Aux-
iliary Output remains synchronized to input signals several dB below the onset of
errors at the DATA Auxiliary Output.
Input and Output The input connectors pass the digitally modulated signal to the receiver mod-
connectors
ule. The input signal, slightly attenuated and available at the OUTPUT connec-
tor, is connected to the input of any of the Agilent 83481,2,3,4,5,6, or 7
modules. The connectors on optical modules include adapters which can eas-
ily be changed to match the type of connectors that are used on your fiber-
optic cables. Refer to “Front-Panel Optical Adapters” on page 4-2 for a
description of the available adapters.
2-5
Operation
Front-Panel Features
Multimode and single-mode connections
Agilent 83492A modules use multimode fiber. Connecting the output to the Optical
INPUT connector on Agilent 83481/2/5 single-mode modules results in large reflections
and insertion loss.
Agilent 83493A modules use 9/125 µm single-mode fiber. Connecting multimode fiber
to the Optical Input connector results in large reflections and insertion loss.
Recovered Clock The recovered clock signal is routed directly to the Agilent 83480A mainframe
through the module’s rear panel. This output has a lower jitter modulation
bandwidth than the front-panel CLOCK Auxiliary Output. Because of the reduced
jitter modulation bandwidth on the mainframe trigger signal, a more complete
view of the jitter on the waveform data is obtained.
2-6
Operation
To Display a Signal
To Display a Signal
1 Install the module as described in “To install the module” on page 1-3. Be sure
to connect all of the cables as described in the procedure.
2 Repeatedly press the SELECT key on the clock recovery module until the front-
panel light indicates the proper data rate of the signal.
Green and red data-rate lights
The data-rate indicator lights change color between red and green to show which data
rate is selected. A red light does not indicate a problem. A red light shows that the adja-
cent red data rate label is selected. A green light shows that the adjacent green data
rate label is selected. Repeatedly pressing the SELECT key cycles through the selections
in one color before switching to the opposite color. On Agilent 83491A modules for
example, the first selection cycle shows 155 Mb/s selected. The second section cycle
shows 1062 Mb/s selected.
• If the UNLOCKED light is on, clock recovery cannot be established on the signal.
• Avoid selecting a data rate that is a multiple of the input signal. For example,
don’t select a 622 Mb/s data rate if the signal is really at 155 Mb/s.
• If you cannot get the clock recovery module to lock on the signal, make sure
that you have selected the correct data rate and that the Agilent 83480A (or
Agilent 54750A) mainframe trigger level is adjusted appropriately.
• Signals displayed using a data trigger are less reliable than using a recovered
clock. Signals triggered on data can also vary depending upon the trigger
level.
2-8
Operation
To Compensate for Module Insertion Loss
To Compensate for Module Insertion Loss
The following steps allow you to enter an offset to compensate for the inser-
tion loss of the clock recovery module. This provides accurate amplitude mea-
surements at the input to the clock recovery module.
1 Disconnect the cable from the clock recovery module’s Input connector.
2 Measure the signal using a power meter. You can use either the
Agilent 83480A’s built-in power meter or an external power meter.
3 Reconnect the cable to the clock recovery module.
4 Disconnect the cable from the reference receiver module’s input connector.
5 Measure the signal using a power meter. You can use either the
Agilent 83480A’s built-in power meter or an external power meter.
Insertion loss:. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ____________
7 On the reference receiver module, press the front-panel channel SETUP key.
8 Press External scale, and set the Atten units to “decibel”.
9 Press Attenuation, and enter the value calculated in Step 6.
2-9
Operation
Using Probes with an Agilent 83491A
Using Probes with an Agilent 83491A
You can use external passive and active probes with the Agilent 83491A elec-
trical clock recovery module. The procedures in this section generate vertical
scale factors. These factors are applied to the calibration of the reference
receiver module’s electrical channel. When selecting a probe, keep in mind
that the input impedance of the Agilent 83491A is 50Ω.
ment to adjust the gain to produce even steps in the vertical scale factors, the
instrument will do so. Typically, probes have standard attenuation factors such
as divide by 10, divide by 20, or divide by 100.
Because the following procedures include compensation for insertion loss of
the clock recovery module, do not perform the procedure “To Compensate for
Module Insertion Loss” on page 2-9.
The following probes are available for use with Agilent 83491A clock recovery
modules:
• Agilent N1020A TDR probe. This passive probe (1:1, 50Ω) provides a fixture
for positioning and holding the probe tip on the device being tested.
• Agilent 54701A 2.5 GHz active probe. This is a 100kΩ, 10:1, probe.
• Agilent 54006A 6 GHz handheld low-impedance probe. This passive probe
(10:1, 500Ω, 20:1, 1kΩ) has an input capacitance of 0.25 pf.
• Agilent 1163A 1 GHz resistive-divider probe. This passive 500Ω probe has an
input capacitance of 1.5 pf.
2-10
Operation
Using Probes with an Agilent 83491A
To compensate for a passive probe
1 Connect the probe to the Input connector on the Agilent 83491A clock recovery
module.
2 Attach the probe tip to the CAL hook that is located near the floppy disk drive.
3 Press the reference receiver module’s front-panel channel SETUP key.
4 Press Calibrate and then Calibrate probe.
To compensate an Agilent 54701A active probe
1 Connect the Agilent 83491A output to the electrical measurement channel
input.
2 Connect the probe to the Input connector on the Agilent 83491A clock recovery
module.
3 Connect the probe power cable to the Probe Power connector on the reference
receiver module.
4 Attach the probe tip to the CAL hook that is located near the floppy disk drive.
5 Press the reference receiver module’s front-panel channel SETUP key.
6 Press Calibrate and then Calibrate probe.
2-11
Operation
Using Probes with an Agilent 83491A
To compensate for other devices
The information in this section applies to both optical and electrical measure-
ments. Since the mainframe’s CAL signal is a voltage source, it cannot be used
to calibrate to the probe tip when the units are set to Ampere, Watt, or
Unknown. Instead, set the external gain and external offset to compensate for
the actual characteristics of the device. If you do not know the actual charac-
teristics, you can refer to the typical specifications that came with the device.
1 Press the reference receiver module’s front-panel channel SETUP key.
2 Press External scale.
3 Press Atten units Ratio, Attenuation 1:1, and then Units Ampere (Volt, Watt, or
Unknown).
4 Press Ext gain, and enter the actual gain characteristics of the device.
5 Press Ext offset, and enter the offset introduced by the device.
2-12
Specifications and Regulatory Information
Specifications and Regulatory Information
Specifications and Regulatory Information
This chapter lists specifications and characteristics of the Agilent 83491/2/3A.
Specifications apply over the temperature range +15°C to +35°C (unless oth-
erwise noted) after the instrument’s temperature has been stabilized after 60
minutes of continuous operation.
Specifications
Characteristics
Specifications described warranted performance.
Characteristics provide useful, nonwarranted, information about the func-
tions and performance of the instrument. Characteristics are printed in
italics.
Calibration cycle Agilent Technologies warrants instrument specifications over the recom-
mended calibration interval. To maintain specifications, periodic recalibrations
are necessary. We recommend that the Agilent 83491/2/3A be calibrated at an
Agilent Technologies service facility every 24 months.
3-2
Specifications and Regulatory Information
Agilent 83491A Specifications
Agilent 83491A Specifications
Table 3-1. Agilent 83491A Specifications
Clock recovery rates (NRZ coding)
155.52 Mb/s
622.08 Mb/s
1062.50 Mb/s
1250 Mb/s
2125.00 Mb/s
2488.32 Mb/s
2500.00 Mb/s
0.1%
0.1%
0.1%
0.1%
0.1%
0.1%
0.1%
Data triggering (characteristic)
50 Mb/s to 2500 Mb/s
Operating input power level a b
Triggering operation, all rates
10 –10 BER, all rates c
–10 dBm to 3 dBm
–10 dBm to 3 dBm
Insertion loss (through path)
DC through 2500 MHz
≤ 7 dB
Output jitter, all rates d
0.0125 UIrms
Maximum continuous electrical power before damage (characteristic)
DATA and CLOCK output amplitude, all rates (characteristic)
1W peak
0.5Vp-p
INPUT electrical return loss
DC through 1250 MHz (characteristic)
1250 MHz through 2500 MHz (characteristic)
≥ 20 dB
≥ 15 dB
DATA and CLOCK electrical return loss
50 MHz through 2000 MHz (characteristic)
2000 MHz through 2500 MHz (characteristic)
≥ 10 dB
≥ 6 dB
a. Source extinction ratio ≥ 8.2 dB when measured per TIA/EIA OFSTP-4A.
b. Operating power level applies over temperature range 25°C 5°.
c. Better than 10 –10 BER when tested with PRBS 223–1 pattern.
d. Measured on an oscilloscope eye diagram with PRBS 223–1 test pattern.
3-3
Specifications and Regulatory Information
Agilent 83492A Specifications
Agilent 83492A Specifications
Table 3-2. Agilent 83492A Specifications (1 of 2)
Wavelength range (characteristic)
750 nm to 860 nm and
1000 nm to 1600 nm
Optical INPUT and OUTPUT fiber (characteristic)
62.5/125 multimode
Optical insertion loss (through path) a
750 nm to 860 nm
1000 nm to 1600 nm
≤ 5.0 dB
≤ 5.0 dB
Optical return loss b
≥ 28 dB
Clock recovery rates (NRZ coding)
155.52 Mb/s
622.08 Mb/s
1062.50 Mb/s
1250 Mb/s
2125.00 Mb/s
2488.32 Mb/s
2500.00 Mb/s
0.1%
0.1%
0.1%
0.1%
0.1%
0.1%
0.1%
Data triggering (characteristic)
50 Mb/s to 2500 Mb/s
d
Operating input power level c
750 nm to 860 nm
–10 dBm to 3 dBm
–10 dBm to 3 dBm
Triggering operation, all rates
10 –10 BER, all rates e
1000 nm to 1600 nm
–13 dBm to 3 dBm
–13 dBm to 3 dBm
Triggering operation, all rates
10 –10 BER, all rates f
Output jitter, all rates g
0.0125 UIrms
Maximum continuous optical power before damage (characteristic)
DATA and CLOCK output amplitude, all rates (characteristic)
10 mW peak
0.5Vp-p
3-4
Specifications and Regulatory Information
Agilent 83492A Specifications
Table 3-2. Agilent 83492A Specifications (2 of 2)
DATA and CLOCK electrical return loss
50 MHz through 2000 MHz (characteristic)
2000 MHz through 2500 MHz (characteristic)
≥ 10 dB
≥ 6 dB
a. Minimum loss in 850 nm window.
b. Single-mode backreflection tested with FC/PC adapter and single-mode fiber. Optical output terminated
with > 33 dB return loss. Return loss with fully filled 62.5 µm core multimode fiber may be slightly lower.
c. Source extinction ratio ≥ 8.2 dB when measured per TIA/EIA OFSTP-4A.
d. Operating power level applies over temperature range 25°C 5°.
e. Better than 10 –10 BER when tested with PRBS 223–1 pattern.
f. Better than 10 –10 BER when tested with PRBS 223–1 pattern.
g. Measured on an oscilloscope eye diagram with PRBS 223–1 test pattern.
3-5
Specifications and Regulatory Information
Agilent 83493A Specifications
Agilent 83493A Specifications
Table 3-3. Agilent 83493A Specifications
Wavelength range (characteristic)
Optical INPUT fiber (characteristic)
Optical insertion loss (through path)
1000 nm to 1600 nm
9/125 single mode
≤ 1.5 dB
Optical return loss a
≥ 28 dB
Clock recovery rates (NRZ coding)
155.52 Mb/s
622.08 Mb/s
1250 Mb/s
2488.32 Mb/s
0.1%
0.1%
0.1%
0.1%
0.1%
2500.00 Mb/s
Data triggering (characteristic)
50 Mb/s to 2500 Mb/s
Operating input power level b c
Triggering operation, all rates
10 –10 BER, all rates d
–20 dBm to 3 dBm
–17 dBm to 3 dBm
Output jitter, all rates e
0.0125 UIrms
Maximum continuous optical power before damage (characteristic)
DATA and CLOCK output amplitude, all rates (characteristic)
10 mW peak
0.5Vp-p
DATA and CLOCK output electrical return loss
50 MHz through 2000 MHz (characteristic)
2000 MHz through 2500 MHz (characteristic)
≥ 10 dB
≥ 6 dB
a. Tested with FC/PC adapter. Optical output terminated without > 33 dB return loss.
b. Source extinction ratio ≥ 8.2 dB when measured per TIA/EIA OFSTP-4A.
c. Operating power level applies over temperature range 25°C 5°.
d. Better than 10 –10 BER when tested with PRBS 223–1 pattern.
e. Measured on an oscilloscope eye diagram with PRBS 223–1 test pattern.
3-6
Specifications and Regulatory Information
Agilent 83491/2/3A Operating Specifications
Agilent 83491/2/3A Operating Specifications
Table 3-4. Agilent 83491/2/3A Operating Specifications
Use
Indoor
Temperature
Operating
Non-operating
0°C to +55°C
–40°C to +70°C
Altitude
Operating
Non-operating
4600 m (15,000 ft)
15,300 m (50,000 ft)
Humidity
Operating
Non-operating
up to 90% relative humidity at <35°C
up to 90% relative humidity at <35°C
Net weight
approximately 1.2 kg (2.6 lb.)
approximately 2.1 kg (4.6 lb.)
Supplied by mainframe
Shipping weight
Power Requirements
3-7
Reference
Front-Panel Optical Adapters
Front-Panel Optical Adapters
Front Panel
Fiber-Optic
Adapter
Description
Agilent Part Number
81000AI
Diamond HMS-10
FC/PCa
D4
81000FI
81000GI
SC
81000KI
DIN
81000SI
ST
81000VI
Biconic
81000WI
Dust Covers
FC connector
1005-0594
1005-0593
1005-0595
1005-0596
1005-0597
Diamond HMS-10 connector
DIN connector
ST connector
SC connector
a. The FC/PC adapter is the standard adapter supplied with the instrument.
4-2
Reference
In Case of Difficulty
In Case of Difficulty
This section provides a list of suggestions for you to follow if the plug-in mod-
ule fails to operate. A list of messages that may be displayed is also included in
this chapter. Before calling Agilent Technologies or returning the unit for ser-
vice, a few minutes spent performing some simple checks may save waiting for
your instrument to be repaired.
If the mainframe does not operate
❒ Is the line fuse good?
❒ Does the line socket have power?
❒ Is the unit plugged in to the proper ac power source?
❒ Is the mainframe turned on?
❒ Is the rear-panel line switch set to on?
❒ Will the mainframe power up without the plug-in module installed?
If the mainframe still does not power up, refer to the optional
Agilent 83480A, Agilent 54750A Service Guide or return the mainframe to a
qualified service department.
If the plug-in does not operate
❒ Is the plug-in module firmly seated in the mainframe slot?
❒ Are the knurled screws at the bottom of the plug-in module finger-tight?
❒ Is the clock recovery module set to the modulation rate of the input signal?
❒ If other equipment, cables, and connectors are being used with the plug-in
module, are they connected properly and operating correctly?
❒ Review the procedure for the test being performed when the problem ap-
4-3
Reference
peared. Are all the settings correct? Can the problem be reproduced?
❒ Are the connectors clean? See “Cleaning Connections for Accurate Measure-
ments” on page 4-10 for more information.
❒ Perform the following procedures:
1 Make sure that the instrument is ready to acquire data by pressing Run.
2 Find any signals on the channel inputs by pressing Autoscale.
3 See if any signals are present at the channel inputs by pressing Trigger, Sweep,
Freerun.
4 After viewing the signal, press triggered.
5 Make sure Channel Display is on by pressing Channel, Display on off, on.
6 Make sure the channel offset is adjusted so the waveform is not clipped off the
display.
7 Make sure the mainframe identifies the plug-in module by pressing Utility, then
System config....
The calibration status of the plug-in modules is listed near the bottom of the
display, in the box labeled “Plug-ins”. If the model number of the plug-
in module is listed next to the appropriate slot number, then the mainframe
has identified the plug-in.
If “~known” is displayed instead of the model number of the plug-in
module, remove and reinsert the plug-in module in the same slot.
If “~known” is still displayed, the mainframe may need to have the latest
operating system firmware installed. Options 001 and 002 provide this
firmware on a 3.5 inch diskette. To load new firmware, follow the
instructions provided with this diskette. If you do not have the optional
diskette, contact your local Agilent Technologies service office (refer to
“Agilent Technologies Service Offices” on page 4-23).
If the mainframe firmware is current and the plug-in module is correctly
installed, then the memory contents of the plug-in module are corrupt.
Contact a qualified service department.
4-4
Reference
Error Messages
Error Messages
The following error messages are for the plug-in module. Typically, the error
messages indicate there is a problem with either the plug-in or the mainframe.
This section explains what the messages mean and offers a few suggestions
that might help resolve the error condition. If the suggestions do not eliminate
the error message, then additional troubleshooting is required that is beyond
the scope of this book. Additional error messages are listed in the
Agilent 83480A, Agilent 54750A User’s Guide for the mainframe.
Memory error occurred in plug-in_:Try reinstalling
plug-in
The mainframe could not correctly read the contents of the memory in the
plug-in.
❒ Remove and reinstall the plug-in module. Each time a plug-in is installed, the
mainframe re-reads the memory in the plug-in module.
❒ Verify the plug-in module is firmly seated in the mainframe slot.
❒ Verify the knurled screws at the bottom of the plug-in module are finger-tight.
❒ Install the plug-in in a different slot in the mainframe.
4-5
Reference
Busy timeout occurred with plug-in_:Try reinstalling plug-in
Busy timeout occurred with plug-in_:Try reinstalling
plug-in
The mainframe is having trouble communicating with the plug-in module.
Make sure there is a good connection between the mainframe and the plug-in
module.
❒ Remove and reinstall the plug-in module.
❒ Verify the plug-in module is firmly seated in the mainframe slot.
❒ Verify the knurled screws at the bottom of the plug-in module are finger-tight.
❒ Install the plug-in in a different slot in the mainframe.
Communications failure exists at slot_:Service is
required
An illegal hardware state is detected at the mainframe-to-plug-in module
interface of the specified slot.
• If the slot is empty, there is a mainframe hardware problem. Refer to the
Agilent 83480A, Agilent 54750A Service Guide.
• If a plug-in is installed in the slot, there is a plug-in module hardware problem.
Return the plug-in module to a qualified service department.
ID error occurred in plug-in_:Service is required
The information read from the memory of the plug-in module does not match
the hardware in the plug-in module. This can be caused by a communication
problem between the mainframe and the plug-in module. Make sure there is a
good connection between the mainframe and the plug-in.
❒ Remove and re-install the plug-in module.
❒ Verify the plug-in module is firmly seated in the mainframe slot.
❒ Verify the knurled screws at the bottom of the plug-in module are finger tight.
❒ The standard Agilent 54750A mainframe does not accept the Agilent 83491/2/
3A module. To use the module, a firmware upgrade must first be installed. Or-
4-6
Reference
Plug-in is not supported:System firmware upgrade is needed
der the Agilent 83480K communications firmware kit and install according to
the instructions.
❒ The Agilent 83480A, Agilent 54750A mainframes do not accept plug-in mod-
ules designed for use with the Agilent 54710A, 54720A.
Plug-in is not supported:System firmware upgrade is
needed
The mainframe may need to have the latest operating system firmware
installed. Options 001 and 002 provide this firmware on a 3.5 inch diskette. To
load the new firmware, follow the instructions provided with the diskette. If
you do not have the optional diskette, contact your local Agilent Technologies
service office.
4-7
Reference
Electrostatic Discharge Information
Electrostatic Discharge Information
Electrostatic discharge (ESD) can damage or destroy electronic components.
All work on electronic assemblies should be performed at a static-safe work
station. The following figure shows an example of a static-safe work station
using two types of ESD protection:
• Conductive table-mat and wrist-strap combination.
• Conductive floor-mat and heel-strap combination.
Figure 4-1. Static-safe work station
4-8
Reference
Electrostatic Discharge Information
Both types, when used together, provide a significant level of ESD protection.
Of the two, only the table-mat and wrist-strap combination provides adequate
ESD protection when used alone.
To ensure user safety, the static-safe accessories must provide at least 1 MΩ of
isolation from ground. Refer to Table 4-1 for information on ordering static-
safe accessories.
W A R N I N G
These techniques for a static-safe work station should not be used
when working on circuitry with a voltage potential greater than 500
volts.
Reducing ESD Damage
The following suggestions may help reduce ESD damage that occurs during
testing and servicing operations.
• Personnel should be grounded with a resistor-isolated wrist strap before re-
moving any assembly from the unit.
• Be sure all instruments are properly earth-grounded to prevent a buildup of
static charge.
Table 4-1. Static-Safe Accessories
Agilent Part
Description
Number
9300-0797
Set includes: 3M static control mat 0.6 m × 1.2 m (2 ft× 4 ft) and 4.6 cm (15
ft) ground wire. (The wrist-strap and wrist-strap cord are not included. They
must be ordered separately.)
9300-0980
9300-1383
Wrist-strap cord 1.5 m (5 ft)
Wrist-strap, color black, stainless steel, without cord, has four adjustable
links and a 7 mm post-type connection.
9300-1169
ESD heel-strap (reusable 6 to 12 months).
4-9
Reference
Cleaning Connections for Accurate Measurements
Cleaning Connections for Accurate
Measurements
Today, advances in measurement capabilities make connectors and connec-
tion techniques more important than ever. Damage to the connectors on cali-
bration and verification devices, test ports, cables, and other devices can
degrade measurement accuracy and damage instruments. Replacing a dam-
aged connector can cost thousands of dollars, not to mention lost time! This
expense can be avoided by observing the simple precautions presented in this
book. This book also contains a brief list of tips for caring for electrical connec-
tors.
Choosing the Right Connector
A critical but often overlooked factor in making a good lightwave measure-
tor types are mainly in the mechanical assembly that holds the ferrule in
position against another identical ferrule. Connectors also vary in the polish,
curve, and concentricity of the core within the cladding. Mating one style of
cable to another requires an adapter. Agilent Technologies offers adapters for
most instruments to allow testing with many different cables. Figure 4-2 on
page 4-11 shows the basic components of a typical connectors.
The system tolerance for reflection and insertion loss must be known when
selecting a connector from the wide variety of currently available connectors.
Some items to consider when selecting a connector are:
• How much insertion loss can be allowed?
• Will the connector need to make multiple connections? Some connectors are
better than others, and some are very poor for making repeated connections.
• What is the reflection tolerance? Can the system take reflection degradation?
• Is an instrument-grade connector with a precision core alignment required?
• Is repeatability tolerance for reflection and loss important? Do your specifica-
4-10
Reference
Cleaning Connections for Accurate Measurements
tions take repeatability uncertainty into account?
• Will a connector degrade the return loss too much, or will a fusion splice be re-
quired? For example, many DFB lasers cannot operate with reflections from
connectors. Often as much as 90 dB isolation is needed.
Figure 4-2. Basic components of a connector.
Over the last few years, the FC/PC style connector has emerged as the most
popular connector for fiber-optic applications. While not the highest perform-
ing connector, it represents a good compromise between performance, reli-
ability, and cost. If properly maintained and cleaned, this connector can
withstand many repeated connections.
However, many instrument specifications require tighter tolerances than most
tolerate connectors with the large non-concentricities of the fiber common
with ceramic style ferrules. When tighter alignment is required, Agilent
Technologies instruments typically use a connector such as the Diamond
HMS-10, which has concentric tolerances within a few tenths of a micron. Agi-
lent Technologies then uses a special universal adapter, which allows other
cable types to mate with this precision connector. See Figure 4-3.
4-11
Reference
Cleaning Connections for Accurate Measurements
Figure 4-3. Universal adapters to Diamond HMS-10.
The HMS-10 encases the fiber within a soft nickel silver (Cu/Ni/Zn) center
which is surrounded by a tough tungsten carbide casing, as shown in
Figure 4-4.
Figure 4-4. Cross-section of the Diamond HMS-10 connector.
The nickel silver allows an active centering process that permits the glass fiber
to be moved to the desired position. This process first stakes the soft nickel
silver to fix the fiber in a near-center location, then uses a post-active staking
to shift the fiber into the desired position within 0.2 µm. This process, plus the
keyed axis, allows very precise core-to-core alignments. This connector is
found on most Agilent Technologies lightwave instruments.
4-12
Reference
Cleaning Connections for Accurate Measurements
The soft core, while allowing precise centering, is also the chief liability of the
connector. The soft material is easily damaged. Care must be taken to mini-
mize excessive scratching and wear. While minor wear is not a problem if the
glass face is not affected, scratches or grit can cause the glass fiber to move
out of alignment. Also, if unkeyed connectors are used, the nickel silver can be
pushed onto the glass surface. Scratches, fiber movement, or glass contamina-
tion will cause loss of signal and increased reflections, resulting in poor return
loss.
Inspecting Connectors
Because fiber-optic connectors are susceptible to damage that is not immedi-
ately obvious to the naked eye, poor measurements result without the user
being aware. Microscopic examination and return loss measurements are the
ensure that optimum connector performance is maintained. With glass-to-
glass interfaces, any degradation of a ferrule or the end of the fiber, any stray
particles, or finger oil can have a significant effect on connector performance.
Where many repeat connections are required, use of a connector saver or
patch cable is recommended.
Figure 4-5 shows the end of a clean fiber-optic cable. The dark circle in the
the light. The surrounding area is the soft nickel-silver ferrule. Figure 4-6
shows a dirty fiber end from neglect or perhaps improper cleaning. Material is
smeared and ground into the end of the fiber causing light scattering and poor
reflection. Not only is the precision polish lost, but this action can grind off the
glass face and destroy the connector.
Figure 4-7 shows physical damage to the glass fiber end caused by either
improper cleaning tools. When severe, the damage of one connector end can
be transferred to another good connector endface that comes in contact with
the damaged one. Periodic checks of fiber ends, and replacing connecting
cables after many connections is a wise practice.
The cure for these problems is disciplined connector care as described in the
following list and in “Cleaning Connectors” on page 4-17.
4-13
Reference
Cleaning Connections for Accurate Measurements
Use the following guidelines to achieve the best possible performance when
making measurements on a fiber-optic system:
• Never use metal or sharp objects to clean a connector and never scrape the
connector.
• Avoid matching gel and oils.
Figure 4-5. Clean, problem-free fiber end and ferrule.
Figure 4-6. Dirty fiber end and ferrule from poor cleaning.
4-14
Reference
Cleaning Connections for Accurate Measurements
Figure 4-7. Damage from improper cleaning.
While these often work well on first insertion, they are great dirt magnets. The
oil or gel grabs and holds grit that is then ground into the end of the fiber.
Also, some early gels were designed for use with the FC, non-contacting con-
nectors, using small glass spheres. When used with contacting connectors,
these glass balls can scratch and pit the fiber. If an index matching gel or oil
must be used, apply it to a freshly cleaned connector, make the measurement,
and then immediately clean it off. Never use a gel for longer-term connections
and never use it to improve a damaged connector. The gel can mask the extent
of damage and continued use of a damaged fiber can transfer damage to the
instrument.
• When inserting a fiber-optic cable into a connector, gently insert it in as
straight a line as possible. Tipping and inserting at an angle can scrape material
off the inside of the connector or even break the inside sleeve of connectors
made with ceramic material.
• When inserting a fiber-optic connector into a connector, make sure that the fi-
ber end does not touch the outside of the mating connector or adapter.
• Avoid over tightening connections.
Unlike common electrical connections, tighter is not better. The purpose of
the connector is to bring two fiber ends together. Once they touch, tightening
only causes a greater force to be applied to the delicate fibers. With connec-
tors that have a convex fiber end, the end can be pushed off-axis resulting in
misalignment and excessive return loss. Many measurements are actually
improved by backing off the connector pressure. Also, if a piece of grit does
happen to get by the cleaning procedure, the tighter connection is more likely
to damage the glass. Tighten the connectors just until the two fibers touch.
4-15
Reference
Cleaning Connections for Accurate Measurements
• Keep connectors covered when not in use.
• Use fusion splices on the more permanent critical nodes. Choose the best con-
nector possible. Replace connecting cables regularly. Frequently measure the
return loss of the connector to check for degradation, and clean every connec-
tor, every time.
All connectors should be treated like the high-quality lens of a good camera.
The weak link in instrument and system reliability is often the inappropriate
use and care of the connector. Because current connectors are so easy to use,
there tends to be reduced vigilance in connector care and cleaning. It takes
only one missed cleaning for a piece of grit to permanently damage the glass
and ruin the connector.
Measuring insertion loss and return loss
Consistent measurements with your lightwave equipment are a good indica-
tion that you have good connections. Since return loss and insertion loss are
key factors in determining optical connector performance they can be used to
determine connector degradation. A smooth, polished fiber end should pro-
duce a good return-loss measurement. The quality of the polish establishes
the difference between the “PC” (physical contact) and the “Super PC” con-
nectors. Most connectors today are physical contact which make glass-to-glass
connections, therefore it is critical that the area around the glass core be clean
and free of scratches. Although the major area of a connector, excluding the
glass, may show scratches and wear, if the glass has maintained its polished
smoothness, the connector can still provide a good low level return loss con-
nection.
If you test your cables and accessories for insertion loss and return loss upon
receipt, and retain the measured data for comparison, you will be able to tell in
the future if any degradation has occurred. Typical values are less than 0.5 dB
of loss, and sometimes as little as 0.1 dB of loss with high performance con-
nectors. Return loss is a measure of reflection: the less reflection the better
(the larger the return loss, the smaller the reflection). The best physically
contacting connectors have return losses better than 50 dB, although 30 to
40 dB is more common.
4-16
Reference
Cleaning Connections for Accurate Measurements
Visual inspection of fiber ends
Visual inspection of fiber ends can be helpful. Contamination or imperfections
on the cable end face can be detected as well as cracks or chips in the fiber
itself. Use a microscope (100X to 200X magnification) to inspect the entire
end face for contamination, raised metal, or dents in the metal as well as any
other imperfections. Inspect the fiber for cracks and chips. Visible imperfec-
tions not touching the fiber core may not affect performance (unless the
imperfections keep the fibers from contacting).
W A R N I N G
Always remove both ends of fiber-optic cables from any instrument,
system, or device before visually inspecting the fiber ends. Disable all
optical sources before disconnecting fiber-optic cables. Failure to do
so may result in permanent injury to your eyes.
Cleaning Connectors
The procedures in this section provide the proper steps for cleaning fiber-
optic cables and Agilent Technologies universal adapters. The initial cleaning,
using the alcohol as a solvent, gently removes any grit and oil. If a caked-on
layer of material is still present, (this can happen if the beryllium-copper sides
of the ferrule retainer get scraped and deposited on the end of the fiber during
insertion of the cable), a second cleaning should be performed. It is not
uncommon for a cable or connector to require more than one cleaning.
C A U T I O N
Agilent Technologies strongly recommends that index matching compounds
not be applied to their instruments and accessories. Some compounds, such as
gels, may be difficult to remove and can contain damaging particulates. If you
think the use of such compounds is necessary, refer to the compound
manufacturer for information on application and cleaning procedures.
Table 4-2. Cleaning Accessories
Item
Agilent Part Number
Any commercially available denatured alcohol
Cotton swabs
—
8520-0023
9300-1223
8500-5262
Small foam swabs
Compressed dust remover (non-residue)
4-17
Reference
Cleaning Connections for Accurate Measurements
Table 4-3. Dust Caps Provided with Lightwave Instruments
Item
Agilent Part Number
Laser shutter cap
FC/PC dust cap
Biconic dust cap
DIN dust cap
08145-64521
08154-44102
08154-44105
5040-9364
HMS10/dust cap
ST dust cap
5040-9361
5040-9366
To clean a non-lensed connector
C A U T I O N
Do not use any type of foam swab to clean optical fiber ends. Foam swabs can
leave filmy deposits on fiber ends that can degrade performance.
1 Apply pure isopropyl alcohol to a clean lint-free cotton swab or lens paper.
Cotton swabs can be used as long as no cotton fibers remain on the fiber end
after cleaning.
2 Clean the ferrules and other parts of the connector while avoiding the end of
the fiber.
3 Apply isopropyl alcohol to a new clean lint-free cotton swab or lens paper.
4 Clean the fiber end with the swab or lens paper.
Do not scrub during this initial cleaning because grit can be caught in the
swab and become a gouging element.
5 Immediately dry the fiber end with a clean, dry, lint-free cotton swab or lens
paper.
6 Blow across the connector end face from a distance of 6 to 8 inches using
filtered, dry, compressed air. Aim the compressed air at a shallow angle to the
fiber end face.
Nitrogen gas or compressed dust remover can also be used.
4-18
Reference
Cleaning Connections for Accurate Measurements
C A U T I O N
Do not shake, tip, or invert compressed air canisters, because this releases
particles in the can into the air. Refer to instructions provided on the
compressed air canister.
7 As soon as the connector is dry, connect or cover it for later use.
If the performance, after the initial cleaning, seems poor try cleaning the con-
nector again. Often a second cleaning will restore proper performance. The
second cleaning should be more arduous with a scrubbing action.
To clean an adapter
The fiber-optic input and output connectors on many Agilent Technologies
instruments employ a universal adapter such as those shown in the following
picture. These adapters allow you to connect the instrument to different types
of fiber-optic cables.
Figure 4-8. Universal adapters.
1 Apply isopropyl alcohol to a clean foam swab.
Cotton swabs can be used as long as no cotton fibers remain after cleaning. The
foam swabs listed in this section’s introduction are small enough to fit into
adapters.
Although foam swabs can leave filmy deposits, these deposits are very thin, and
the risk of other contamination buildup on the inside of adapters greatly out-
weighs the risk of contamination by foam swabs.
2 Clean the adapter with the foam swab.
3 Dry the inside of the adapter with a clean, dry, foam swab.
4 Blow through the adapter using filtered, dry, compressed air.
Nitrogen gas or compressed dust remover can also be used. Do not shake, tip,
or invert compressed air canisters, because this releases particles in the can
into the air. Refer to instructions provided on the compressed air canister.
4-19
Reference
Returning the Instrument for Service
Returning the Instrument for Service
ment for repair or calibration. Always call the Agilent Technologies Instrument
Support Center first to initiate service before returning your instrument to a
service office. This ensures that the repair (or calibration) can be properly
tracked and that your instrument will be returned to you as quickly as possi-
ble. Call this number regardless of where you are located. Refer to “Agilent
Technologies Service Offices” on page 4-23 for a list of service offices.
Agilent Technologies Instrument Support Center . . . . . . . . . . .(800) 403-0801
If the instrument is still under warranty or is covered by an Agilent Technolo-
gies maintenance contract, it will be repaired under the terms of the warranty
or contract (the warranty is at the front of this manual). If the instrument is
no longer under warranty or is not covered by an Agilent Technologies mainte-
nance plan, Agilent Technologies will notify you of the cost of the repair after
examining the unit.
When an instrument is returned to a Agilent Technologies service office for
servicing, it must be adequately packaged and have a complete description of
the failure symptoms attached. When describing the failure, please be as spe-
cific as possible about the nature of the problem. Include copies of additional
failure information (such as the instrument failure settings, data related to
instrument failure, and error messages) along with the instrument being
returned.
Preparing the instrument for shipping
1 Write a complete description of the failure and attach it to the instrument.
Include any specific performance details related to the problem. The following
4-20
Reference
Returning the Instrument for Service
information should be returned with the instrument.
• Type of service required.
• Date instrument was returned for repair.
• Description of the problem:
• Whether problem is constant or intermittent.
• Whether instrument is temperature-sensitive.
• Whether instrument is vibration-sensitive.
• Instrument settings required to reproduce the problem.
• Performance data.
• Company name and return address.
• Name and phone number of technical contact person.
• Model number of returned instrument.
• Full serial number of returned instrument.
• List of any accessories returned with instrument.
2 Cover all front or rear-panel connectors that were originally covered when you
first received the instrument.
C A U T I O N
C A U T I O N
Cover electrical connectors to protect sensitive components from electrostatic
damage. Cover optical connectors to protect them from damage due to physical
contact or dust.
Instrument damage can result from using packaging materials other than the
original materials. Never use styrene pellets as packaging material. They do not
adequately cushion the instrument or prevent it from shifting in the carton.
They may also cause instrument damage by generating static electricity.
3 Pack the instrument in the original shipping containers. Original materials are
available through any Agilent Technologies office. Or, use the following
guidelines:
• Wrap the instrument in antistatic plastic to reduce the possibility of damage
caused by electrostatic discharge.
• For instruments weighing less than 54 kg (120 lb), use a double-walled, cor-
rugated cardboard carton of 159 kg (350 lb) test strength.
• The carton must be large enough to allow approximately 7 cm (3 inches) on
all sides of the instrument for packing material, and strong enough to accom-
modate the weight of the instrument.
• Surround the equipment with approximately 7 cm (3 inches) of packing ma-
terial, to protect the instrument and prevent it from moving in the carton. If
packing foam is not available, the best alternative is S.D-240 Air Cap™ from
4-21
Reference
Returning the Instrument for Service
Sealed Air Corporation (Commerce, California 90001). Air Cap looks like a
plastic sheet filled with air bubbles. Use the pink (antistatic) Air Cap™ to
reduce static electricity. Wrapping the instrument several times in this ma-
terial will protect the instrument and prevent it from moving in the carton.
4 Seal the carton with strong nylon adhesive tape.
5 Mark the carton “FRAGILE, HANDLE WITH CARE”.
6 Retain copies of all shipping papers.
4-22
Reference
Agilent Technologies Service Offices
Agilent Technologies Service Offices
Before returning an instrument for service, call the Agilent Technologies
Instrument Support Center at (800) 403-0801, visit the Test and Measurement
index.html, or call one of the numbers listed below.
Agilent Technologies Service Numbers
Austria
01/25125-7171
32-2-778.37.71
(11) 7297-8600
86 10 6261 3819
45 99 12 88
Belgium
Brazil
China
Denmark
Finland
358-10-855-2360
01.69.82.66.66
0180/524-6330
080-34 35788
+39 02 9212 2701
01 615 8222
France
Germany
India
Italy
Ireland
Japan
(81)-426-56-7832
82/2-3770-0419
(5) 258-4826
Korea
Mexico
Netherlands
Norway
Russia
020-547 6463
22 73 57 59
+7-095-797-3930
(34/91) 631 1213
08-5064 8700
(01) 735 7200
01 344 366666
(800) 403-0801
Spain
Sweden
Switzerland
United Kingdom
United States/Canada
4-23
Index
DATA Auxiliary Output connector, 2-5
data rate
declaration of conformity, 3-8
dust caps, 4-18
A
adapter cable, 1-3, 1-4
Agilent 54701A active probe, 2-11
Agilent 83492A
connections, 1-7
maximum input level, 1-5
Agilent offices, 4-23
B
electrostatic discharge, 1-2, 4-8
Bellcore GR-253-CORE, 2-5
block diagram, 2-7
bypass mode, 2-4, 2-5
ESD
C
cabinet, cleaning, iii
calibration cycle, 3-2
care
of cabinet, iii
care of fiber optics, 1-2
channel
key, 1-2
setup, 1-2
Channel key, 2-3
characteristics, defined, 3-2
classification
product, iii
cleaning
fiber
single-mode, 2-2
fiber optics
upgrade kit, 1-2
adapters, 4-2
adapters, 4-19
cabinet, iii
fiber-optic connections, 4-10, 4-18
non-lensed connectors, 4-18
compensation
insertion loss, 2-9
other devices, 2-12
passive probe, 2-11
compressed dust remover, 4-17
connector
features, 2-4
fuse
care, 4-10
cotton swabs, 4-17
customer assistance, iv
G
GPIB programming, 2-3
green light, 1-5
D
damaged shipment, 1-3
Index-1
Index
Agilent 83492A, 3-4
Agilent 83493A, 3-6
defined, 3-2
swabs, 4-17
I
IEC Publication 61010-1, iii
input
connector, 4-10
Input connector, 1-2, 1-7
input connector, 2-2
input signal, maximum safe, 1-5
insertion loss, 2-3, 2-6
compensation, 2-9
instrument
returning for service, 4-20
ITU-T G.957, 2-5
technical assistance, iv
testing, responsivity, 1-4
trigger bypass mode, 2-4, 2-5
troubleshooting, 4-3
M
mainframe troubleshooting, 4-3
maximum input level, 1-5
O
Output connector, 2-5
Unlocked light, 1-6, 2-5, 2-8
P
packaging for shipment, 4-21
plug-in module
serial number, 1-3
programming, 2-3
R
red light, 1-5
regulatory information, 3-2
responsivity, testing, 1-4
S
safety, iii
laser classification, iii
safety information, iv, 1-3, 2-3
sales and service offices, 4-23
SELECT key, 1-5, 2-4
serial number, 1-3
service, 4-20
returning for, 4-20
sales and service offices, 4-23
shipping
Index-2
|