M2922A, M2925A, M2926A
Philips FM-2
Portable Antepartum
Fetal Monitor
• 12-hours memory for fetal traces.
Features
• Internal battery power or AC power from external
power supply.
• 6-hours (approx.) battery back-up operation,
rechargeable during normal monitoring.
• Download capability to transmit traces stored in
its memory to a PC or a Philips OB TraceVue
surveillance and alerting system.
• Online transmission to a monitoring system such
as OB TraceVue.
• Easily connects to off-the-shelf printer for
documentation.
Description
FM-2 is a fetal monitor for non invasively
measuring, and showing maternal uterine
contractions and the fetal heart rate as a graphic
display, and optionally on a strip chart recorder,
printer or OB TraceVue system. The data is
intended to aid in assessing the well being of the
fetus during the final trimester of pregnancy non-
stress test (NST).
• Lightweight and portable, with carrying handle.
• Easy to read display for viewing of FHR and Toco
traces and numerics.
• On-screen scrolling for viewing stored fetal traces.
• 15 BPM separated grid lines for easy viewing of
fetal heart rate accelerations for NST assessment.
This portable fetal monitor offers a solution for
antepartum screening in the private doctor’s office,
outpatient clinics, and for remote applications such
as traveling nurses visiting patient’s homes, or for
clinics in rural areas seeking remote consultation.
• Twins ultrasound monitoring capability.
• Configurable audible and flashing fetal heart rate
alarms.
• Uses watertight transducers compatible with entire
Series 50 range of fetal monitors.
It is intended for use only by trained medical
personnel located in hospitals, clinics, doctor’s
offices and in the patient’s home or during transport
under the direction of a licensed practitioner.
• Event marker for easy documentation of events
and kick counts.
• M1913A - Paper
Monitoring Accessories
Case of 40 packs thermal recorder paper, 0-100
mmHg IUP scale, 50-210 FHR scale, 1 min
(3cm) highlighted lines. Green. Not available in
Europe.
Standard Accessories and Supplies
These are included with the monitor
• 1 x Doppler Ultrasound FHR Transducer-
1
M1356A
Associated Products
• 1 x External TOCO Transducer - M1355A
• 1 x Remote Event Marker - 15249A
External Recorder (M2925A)
The recorder is an external device for printing stored
fetal traces, or realtime fetal trace output. It connects
to the FM-2 via a standard (IEEE 1284 compliant))
printer cable.
• 2 x Reusable Abdominal Belts - M1562A
• 1 pack of three transducer knob adapters -
M1356-43203
FM-2 Viewer Utility Software (M2926A)
• 1x 250ml ultrasound gel
The monitor can send data to an IBM-compatible
PC running Microsoft NT4 or 98.
• User’s Guide
• Pocket Guide to Fetal Monitoring and Assessment
by Susan Martin Tucker (English language
shipments only). ISBN 0-323-00884-4,
4th edition
See the specifications section of this data sheet for
system requirements.
Mounting (M3904A and M3909A)
• M3904A - GCX Wallmounting.
• M3909A - GCX Rollstand
• Service Guide
Additional Supplies and Accessories
These are not included with the monitor, but can be
purchased separately.
• M1356A - Second Doppler Ultrasound FHR
Transducer for twins monitoring.
• M3919A - Canvas Tote Bag
with room for monitor plus accessories.
• M1569A - Belt Fastening Button
Pack of 10.
• M2208A - 100 Disposable Abdominal Belts
1.3m long, 50mm wide (52” long, 2” wide). 1 box
of 50 blue and 50 pink.
• 40483A - Ultrasound Transmission Gel
12 x 250 ml bottles, shelf life 24 months
maximum, 6 months minimum.
• 40483A - 5 Liter Refill Container
shelf life 24 months maximum, 6 months
minimum.
• M1910A - Paper
Case of 40 packs thermal recorder paper, 0-100
mmHg IUP scale, 30-240 FHR scale, 1 min
(3cm) highlights. Orange (USA).
• M1911A - Paper
Case of 40 packs thermal recorder paper, 0-100
mmHg IUP scale, plus 0-13 kPa scale, 50-210
FHR scale, no highlighted 3 cm lines. Green
(International).
1.The monitor has twins capability built in. To monitor
twins you require an additional ultrasound transducer.
Purchase additional M1356A.
3
Recharge without
monitoring:
Maximum 8 hours to
full recharge when no
monitoring occurs (typical
4.75 hours).
Specifications
Monitor (M2922A)
Federal Law restricts this device to use by or on the
Low power warning: Approximately 30 minutes
1
at 25oC.
order of a physician.
Power - External Power Supply
Standards Com pliance
AC-powered
110-120V~, 50-60Hz,
0.15A
or
220-240V~, 50-60Hz,
0.08A
The FM-2 Portable Antepartum Fetal Monitor
complies with the requirements of the Medical
Device Directive 93/42/EEC and the following
safety standards:
• EN60601-1/IEC60601-1, EN60601-1-1,
EN60101-2-37
Display
• IEC 1157
• IEC 60601-1
• UL2601-1
• CAN/CSA C22.2#601.1-M90
Backlight:
Active area:
Cold cathode fluorescent
3.78” x 2.83”
(320 x 240 dots)
85 dots per inch
10:1
Resolution:
Contrast:
Class II equipment, double insulated
Continuous operation
Ultrasound
Isata
BPM range:
Paper scale:
:
< 4mW
Input
Connector
Resulting
Isolation
Param eter
USA paper: 30-240 BPM,
International Paper: 50-210
BPM.
US
Toco
BF
BF
BF
BF
Ultrasound
processing range: 50-240 BPM.
Electrom agnetic Com patibility
Accuracy:
±1% ±1 BPM.
Leakage current:
≤10 µ A @ 264 VAC applied
to transducer.
Complies with EN60601-1-2
Isolation:
>4 kV RMS, Type BF
applied part.
Environm ental
Operating Temp:
Storage Temp:
-10oC to 45oC
Toco
(50oF to 110oF)
-20oC to 60oC
(-29oF to 175oF)
Range:
0-100 relative units
1 count
±1% ±1 relative unit
≤10 µ A @ 264 VAC applied
to transducer
>4 kV RMS, Type BF
applied part
Resolution:
Accuracy:
Leakage:
Relative Humidity: 20% to 90% non
condensing
0-3048m (10,000 ft)
Altitude:
Isolation:
Dim ensions
Height:
Width:
Depth:
Weight:
15.24 cm (6.0 inches)
26.4 cm (10.4 inches)
13.7 cm (5.4 inches)
2.27 kg (5lb)
Transducers
Doppler Ultrasound FHR (M1356A)
Weight:
185 g (6.5 oz)
2.5m (8ft 2in)
75mm (2.95in)
Power - Battery
Cable length:
Diameter:
Latex free
Type:
Operation:
Lead acid, rechargeable.
minimum 2-hours fully
o
charged at 10 C, longer
External Toco (M1355A)
at higher temperature.
Maximum 14 hours full
recharge during monitor-
ing (typical 11 hours).
Transducer Weight: 180 g (6.3 oz)
Cable length: 2.5m (8ft 2in)
Diameter: 75mm (2.95in)
Latex free
Recharge during
monitoring:
Rem ote Event Marker (15249A)
Length: 2.8m (9ft 2in)
1.USA only
4
Weight 75g (2.65oz)
Latex free
Width:
Depth:
Weight:
23.4cm (9.2 inches)
26.2 cm (10.3 inches)
3.54 kg (7lb 13oz)
Reusable Abdom inal Belt (M1562A)
Standards Com pliance
Length: 1.3m (52in)
Width: 50mm (2in)
Safety
With button holes, and pre-attached buttons. Can
be further cut. Contains latex.
- Complies with EN60601-1, EN60601-1-1
- Class I equipment, requires 3-prong outlet with
protective earth
FM-2 Viewer Utility Software
(M2926A)
- Continuous operation
Electrom agnetic Com patibility
Windows operating system-based utility software.
Supplied on CD-ROM, together with operating
instructions, license and 3m PC connection cable.
Complies with EN60601-1-2
Power
System requirements:
AC-powered
110-120V~, 50-60Hz,
0.15A
or
• IBM-compatible PC (200MHz Pentium or
equivalent)
220-240V~, 50-60Hz,
0.08A
• Standard VGA graphics adapter with 8MB
memory
Consumption:
20 watts maximum at
normal supply voltage
• CD-ROM drive
• 60 MB (minimum) available space on hard drive
Environm ental
• Windows 98, or NT (with service pack 6a)
-10oC to 45oC
(50oF to 110oF)
-20oC to 60oC
(-29oF to 175oF)
operating system
Operating Temp:
Microsoft®, and Windows® are either registered
trademarks or trademarks of Microsoft Corporation
in the United States and/or other countries.
Storage Temp:
Relative Humidity: 20 to 90% non condensing
Altitude: 0-3048m (10,000ft)
The FM-2 viewer utility software complies with the
requirements of the Medical Device Directive 93/
42/EEC.
Recorder Paper (Therm al)
Federal Law restricts this device to use by or on the
order of a physician.
Style:
Z-fold
Pack size:
End-of-pack:
150 pages
Mark along paper edge,
stop sign on last page
Front door, drop-in, self
aligning
Printers
Supported printers are:
• HP DeskJet 990Cxi
• HP DeskJet 980Cxi
• HP DeskJet 960c
• HP DeskJet 840c
• HP DeskJet 640c
• HP DeskJet 350
• HP Photosmart 1215
Loading:
Paper detectors:
Paper speeds:
Paper out, and recorder
loading door open
Normal (1, 2, and 3 cm/
min ±1%) and batch print
(average 25 cm/min)
±1% (exclusive of paper
accuracy)
Paper tracking
accuracy:
M1910A:
30-240 FHR range, orange
(Canada and USA) grid, no kPa scale, 3cm
highlighted lines.
This list will be updated periodically. Contact
Philips Medical sales for details about currently
supported printers.
M1911A:
(International)
50-210 FHR range, green
grid, kPa scale, no
highlighted lines.
M1913A/J:
(Japan)
50-210 FHR range, green
grid, no kPa scale, 3 cm
highlighted lines.
Recorder (M2925A)
Dim ensions
Height:
12.5 cm (4.9 inches)
5
Philips offers a wide range of fetal monitors for
different environments and applications. Review the
capabilities of all fetal monitors to help you choose
the monitor that is right for your needs.
Feature
Table legend:
•
Standard
o O ptional
−
Unavailable
Designed for portability/transport
Integrated fetal trace display
On screen NST assessment aid
Battery operation and AC operation from power supply
Off the shelf printer interface
PC utility software
•
•
•
•
•
•
•
−
•
•
−
−
−
−
−
−
−
•
−
−
−
−
−
−
−
•
−
−
−
−
−
−
−
•
−
−
−
−
−
−
−
•
External recorder
Integrated recorder
Fetal trace data storage (12 hours)
−
o
−
−
−
−
−
−
Transmission to obstetrical surveillance system (for example
OB TraceVue) via phone line
Twins heart rate separation
•
•
•
•
•
•
•
•
•
•
Direct system interface to obstetrical overview system such
as OB TraceVue
Watertight, easy to clean transducers
•
−
•
•
•
•
•
•
•
•
Indication when two channels record the same heart rate
(Cross Channel Verification - CCV)
Indication when you accidentally record maternal SpO
−
−
−
−
•
2
(Cross Channel Verification plus - CCV+)
Performance PLUS! technology to ensure continuous,
accurate and high-quality ultrasound traces
−
−
•
•
•
•
•
•
•
•
Precision signal track and hold to reduce artifacts and
increase sensitivity
Interface to external fetal pulse oximeter (FSpO )
−
−
o
o
o
o
o
−
•
2
Gross fetal body movement recording (fetal movement
profile - FMP)
•
External fetal measurements
Single US FHR
Twin US FHR
Toco
•
•
•
−
−
−
−
−
−
•
•
•
•
•
•
−
−
•
•
•
•
•
•
•
•
•
•
•
•
o
•
Internal fetal measurements
DECG
−
−
−
−
−
o
IUP/TTUIP
Fetal SpO (FSpO )
−
2
2
Integrated maternal
measurements
NIBP and SpO
•
•
o
•
•
2
MECG
ob
Interface to external maternal monitor
o
a.Series 50 XMO not available in USA or Japan
b. When maternal parameters are not already integrated in the Series 50 XM/XMO monitor (that is, you have not ordered option A03)
6
Philips Medical Systems is part of Royal Philips Electronics
CO RPO RATE HEADQ UARTERS:
Philips Medical Systems
Nederland B.V.
Postbus 10.000
5680 DA Best
Netherlands
UNITED STATES:
Philips Medical Systems
Cardiac and Monitoring Systems
3000 Minuteman Road
Andover, MA 01810
(800) 934-7372
CANADA:
Philips Medical Systems
2660 Matheson Blvd. E.
Mississauga, Ontario L4W 5M2
(800) 291-6743
EURO PE, MIDDLE EAST AND AFRICA:
Philips Medizinsysteme Böblingen GmbH
Cardiac and Monitoring Systems
Hewlett-Packard Str. 2
71034 Böblingen
Germany
Fax: (+49) 7031 463 1552
LATIN AMERICA HEADQ UARTERS:
Philips Medical Systems
5200 Blue Lagoon Drive
9th Floor
Miami, FL 33126
(305) 267-4220
ASIA PACIFIC HEADQ UARTERS:
Philips Medical Systems
24F Cityplaza One
1111 King's Road
Taikoo Shing, Hong Kong
(+852) 3197 7777
© 2002 Philips Medizinsysteme Böblingen GmbH
All rights are reserved. Reproduction in whole or in part is
prohibited without the prior written consent of the copyright
holder.
Published April, 2002
5990-0507EN
|