Omega Engineering Barcode Reader Om Daq Usb 2401 User Manual

®
Users Guide  
Shop online at  
®
omega.com  
For latest product manuals:  
omegamanual.info  
OM-DAQ-USB-2401  
Multiple Channel USB Data  
Acquisition Module  
OM-DAQ-USB-2400  
Multiple Channel USB Data Acquisition Module  
TABLE OF  
CONTENTS  
Page  
Section 1 - Introduction ....................................................................... 1-1  
1.1 Precautions ................................................................................................ 1-1  
1.2 Safety Warnings and IEC Symbols ......................................................... 1-2  
1.3 Statement on CE Marking ........................................................................ 1-2  
1.4 General Description and System Components .................................... 1-2  
Section 2 - Hardware .......................................................................... 2-1  
2.1 Package Inspection ................................................................................... 2-1  
2.2 Included Items .......................................................................................... 2-1  
Section 3 - Hardware Setup ................................................................. 3-1  
3.1 OM-DAQ-USB-2400 Components ......................................................... 3-1  
3.2 USB Hubs and Power Adaptors ............................................................. 3-1  
3.2.1 USB-powered Hubs .................................................................... 3-2  
3.2.2 Self-powered Hubs ..................................................................... 3-2  
3.2.3 Power Adaptors .......................................................................... 3-2  
3.3 Connecting Your OM-DAQ-USB-2401 Acquistition Systems To PC ... 3-3  
3.4 Connecting Various Hardware Setups .................................................. 3-4  
3.4.1 Direct Connection To Computer USB Port(s) ......................... 3-4  
3.4.2 Connection To USB-powered Hub ........................................... 3-5  
3.4.3 Connections To Self-powered and USB-powered Hubs ........ 3-5  
3.5 OM-DAQ-USB-2401 Mounting .............................................................. 3-6  
3.5.1 OM-DAQ-USB-2401 Wall Mounting ........................................ 3-7  
3.5.2 OM-DAQ-USB-2401 DIN Rail Mounting ................................ 3-8  
Section 4 - Software ............................................................................ 4-1  
4.1 Getting Started .......................................................................................... 4-1  
4.2 Software Installation ................................................................................ 4-1  
4.2.1 System Requirements ................................................................. 4-1  
4.2.2 Software Setup ............................................................................. 4-2  
4.3 DAQ Central Software Operation .......................................................... 4-4  
4.3.1 Main Control Window Pull-Down Operations ...................... 4-6  
4.3.2 Device Configuration ................................................................. 4-8  
4.4 Power Line Rejection .............................................................................. 4-14  
4.5 Optical Isolation ....................................................................................... 4-14  
4.6 A/D Conversion ..................................................................................... 4-14  
4.7 Input Ranges .................................................................................................... 4-15  
4.8 Analog Input Configuration .................................................................. 4-15  
4.9 Measurement Duration, Scan Time, and Resolution ......................... 4-16  
4.10 Automatic Calibration ...........................................................................4-17  
4.11 Thermocouple Measurements ............................................................ 4-18  
Section 5 - Signal Management ............................................................ 5-1  
5.1 Channel Control and Expansion ............................................................ 5-1  
5.2 Scan Time and Resolution ....................................................................... 5-1  
5.2.1 Scan Time ..................................................................................... 5-1  
5.2.2 Resolution [Effective Number of Bits (ENOB, RMS)] .............5-1  
5.3 Under Sampling and Aliasing ................................................................ 5-2  
i
OM-DAQ-USB-2400  
Multiple Channel USB Data Acquisition Module  
TABLE OF  
CONTENTS  
Page  
5.4 Triggering ................................................................................................... 5-4  
5.5 Signal Modes ............................................................................................. 5-4  
5.6 System Noise ............................................................................................. 5-6  
5.7 Averaging ................................................................................................... 5-6  
5.8 Analog Filtering ........................................................................................ 5-6  
5.9 Input and Source Impedance .................................................................. 5-7  
5.10 Crosstalk .................................................................................................. 5-7  
8.1 Service and Calibration ............................................................................ 7-1  
Section 6 - CE Conformity ..................................................................... 6-1  
6.1 OM-DAQ-USB-2401 Design for CE Conformity .................................. 6-1  
Section 7 - Troubleshooting .................................................................. 7-1  
7.1 Basic Checklist ........................................................................................... 7-1  
7.2 Symptoms and Solutions ......................................................................... 7-1  
Section 8 - Service and Calibration ....................................................... 8-1  
8.1 Service and Calibration ............................................................................ 8-1  
Section 9 - Specifications ..................................................................... 9-1  
Section 10 - Approvals and Regulatory Compliance ........................... 10-1  
10.1 International Usage and CE Marking ................................................ 10-1  
ii  
OM-DAQ-USB-2400  
Multiple Channel USB Data Acquisition Module  
LIST OF  
FIGURES  
List of Figures  
Section  
Figure  
1-1  
Description .............................................................. Page  
IEC Symbols ............................................................... 1-2  
System Components ................................................. 1-3  
Section 1  
1-2  
Section 3  
3-1  
3-2  
OM-DAQ-USB-2401 Hardware and Dimensions ... 3-1  
Connection of an OM-DAQ-USB-2401 System  
to a Computer USB Port ........................................... 3-3  
3-3  
3-4  
3-5  
Direct Connection to Computer USB Port ............. 3-4  
Connection to USB-powered Hub .......................... 3-5  
Connection to Self-powered Hub  
and to USB-powered Hub ........................................ 3-6  
3-6  
3-7  
OM-DAQ-USB-2401 Wall Mounting ...................... 3-7  
OM-DAQ-USB-2401 Wall Mounting  
Dimensions ................................................................ 3-7  
3-8  
OM-DAQ-USB-2401 Rail Mounting ....................... 3-8  
Section 4  
4-1  
4-2  
4-3  
Welcome Screen ......................................................... 4-2  
Select Install Screen ................................................... 4-2  
Confirm Installation Screen ..................................... 4-3  
Main Control Window ............................................. 4-4  
Toolbar Buttons ......................................................... 4-5  
File Menu .................................................................... 4-6  
Device Menu .............................................................. 4-6  
Tools Menu ................................................................. 4-7  
Help Menu ................................................................. 4-7  
Configuration Mode Channels Tab Screen ............ 4-8  
Scan Options .............................................................. 4-9  
Unit Options .............................................................. 4-9  
Automatic Calibration .............................................. 4-9  
Trigger Options ....................................................... 4-10  
"Start Trigger" Options ............................................ 4-10  
"Stop Trigger" Options ........................................... 4-10  
Data Tab .................................................................... 4-11  
Data "Save Behavior" Option ................................. 4-11  
Main Window Screen ............................................. 4-12  
Gauges Screen .......................................................... 4-12  
Bars Screen ............................................................... 4-12  
Digital Screen ........................................................... 4-12  
Table Screen .............................................................. 4-12  
Waveform Screen .................................................... 4-13  
Sample Options Screen ........................................... 4-13  
4-4  
Table 4-1  
4-5  
4-6  
4-7  
4-8  
4-9  
4-10  
4-11  
4-12  
4-13  
4-14  
4-15  
4-16  
4-17  
4-18  
4-19  
4-20  
4-21  
4-22  
4-23  
4-24  
iii  
OM-DAQ-USB-2400  
Multiple Channel USB Data Acquisition Module  
LIST OF  
FIGURES  
List of Figures  
Section  
Figure  
4-25  
Description ............................................................. Page  
OM-DAQ-USB-2400 Block Diagram .................... 4-14  
4-26  
OM-DAQ-USB-2401 Single-ended and Differential  
Conections to Analog Input Channels Diagram ... 4-16  
4-27  
4-28  
Calibration/Scan Arrangement ............................ 4-17  
Thermocouple Wiring ............................................ 4-18  
Section 5 Table 5-1  
Scan Rate vs Resolution (ENOB, RMS) .................. 5-2  
Examples of Under Sampling ................................. 5-3  
Trigger Source Diagram ........................................... 5-4  
Noise Reduction in Differential Mode ................... 5-5  
Example of Floating Differential Circuit .................5-5  
Analog to Digital Converter .................................... 5-7  
5-1  
5-2  
5-3  
5-4  
5-5  
Section 6  
6-1  
6-2  
OM-DAQ-USB-2401 With Ferrite Cores Installed .. 6-1  
Ferrite Core ................................................................ 6-2  
Section 9 Table 9-1  
Thermocouple Accuracy (C) for  
OM-DAQ-USB-2400 .................................................. 9-1  
iv  
1
Introduction  
Section 1 - Introduction  
Please read this manual completely before installing and operating your Omega  
data acquisition system. It’s important to read and follow all notes, cautions,  
warnings and safety precautions before operating this device. “Device” refers to  
your data acquisition unit.  
1.1 Precautions  
• This device is not designed for use in any medical or nuclear applications.  
• Do not operate this device in flammable or explosive environments.  
• Never operate with a power source other than the one recommended in this  
manual or listed on product labels.  
• This device has been designed for dry, moisture free indoor applications only.  
• Do not operate this device outside of the recommended use outlined in this  
manual.  
• Never use your data acquisition unit as a portable device. Your unit has been  
designed to be operated in a permanent installation only.  
NOTE:  
There are no user serviceable parts inside your device.  
Attempting to repair or service your unit may void your  
warranty.  
NOTE:  
The discharge of static electricity can damage some electronic  
components. Semiconductor devices are especially susceptible  
to ESD damage. You should always handle components  
carefully, and you should never touch connector pins or  
circuit components unless you are following ESD guidelines in  
an appropriate ESD controlled area. Such guidelines include  
the use of properly grounded mats and wrist straps, ESD  
bags and cartons, and related procedures.  
NOTE:  
Never remove a USB cable from an active OM-DAQ-USB-  
2401 unit while an acquisition is in progress. An active unit is  
any unit that is currently open and has channels configured  
for scanned input. Such disconnection may require you to exit  
and then re-launch Omega DAQ Central, after the USB cable  
has been connected.  
1-1  
1
Introduction  
NOTE:  
When using OM-DAQ-USB-2401 unit to acquire data,  
computer energy save modes can cause false data readings.  
Prior to using the unit, ensure your computers energy save  
mode is disabled. If needed, consult your PC users manual to  
disable energy save (power suspension) modes.  
1.2 Safety Warnings and IEC Symbols  
This device is marked with international safety and hazard symbols in  
accordance with IEC standards. It is important to read and follow all precautions  
and instructions in this manual before operating or commissioning this device as  
it contains important information relating to safety and EMC. Failure to follow  
all safety precautions may result in injury and or damage to your device. Use of  
this device in a manner not specified will void your warranty.  
IEC symbols  
Description  
Caution, refer to accompanying documentation  
EU’s Waste Electrical and Electronic Equipment  
Compliance  
Figure 1-1. IEC Symbols  
1.3 Statement on CE Marking  
®
It is the policy of OMEGA to comply with all worldwide safety and EMI/EMC  
regulations that apply. OMEGA is constantly pursuing certification of its  
products to the European New Approach Directives. OMEGA will add the CE  
mark to every appropriate device upon certification. For additional information  
see Section 10 - Approvals & Regulatory Compliance.  
1.4 General Description & System Components  
The OM-DAQ-USB-2401 is a USB 2.0 full speed thermocouple/voltage input  
data acquisition module (fully compatible with both USB 1.1 and USB 2.0 ports).  
This stand-alone module draws power from the USB port to operate. An external  
(optional) power supply can be used. All configurable options (including  
individual channel input type and range) are software programmable.  
The OM-DAQ-USB-2401 has user programmable voltage inputs that range from  
30 mV to 10V, full scale. The compact, modular packaging ensures ease of use  
in a variety of applications. Units can be DIN rail or wall mounted with the  
included hardware or easily operated on a bench. All analog input channels can  
be measured sequentially at about 1 ms per channel. A total of 1000 samples per  
second can be taken, divided across all active channels.*  
1-2  
1
Introduction  
NOTE:  
*At highest scan rate, 1000 samples/sec +/- 1% with one  
channel on, +/- 5% with all channels on.  
The main items you will be using in your data acquisition system are the  
OM-DAQ-USB-2401 unit, its USB cable, and power adaptor.  
POWER  
ADAPTER  
(OPTIONAL)  
OM-DAQ-USB-2401  
MODULE  
USB  
PC WITH USB CONNECTION (NOT INCLUDED)  
Figure 1-2. System Components  
1-3  
2
Hardware  
Section 2 - Hardware  
It is important that you read this manual completely and follow all safety  
precautions before operating this instrument.  
2.1 Package Inspection  
Remove the packing list and verify that you have received all your equipment. If  
you have any questions about the shipment, please call our Customer Service  
Department at 1-800-622-2378 or 203-359-1660. We can also be reached on the  
Internet at omega.com, e-mail: [email protected]. When you receive the  
shipment, inspect the container and equipment for any signs of damage. Note  
any evidence of rough handling in transit. Immediately report any damage to  
the shipping agent.  
NOTE:  
The carrier will not honor any damage claims unless all  
shipping material is saved for inspection. After examining  
and removing contents, save packing material and carton in  
the event reshipment is necessary.  
2.2 Included Items  
The following items are supplied in the box:  
• 1 Omega OM-DAQ-USB-2401 data acquisition module  
• 1 Omega DAQ Central software CD  
• 1 USB Cable  
• 1 9Vdc Power Adaptor  
• 1 Mounting Kit (8 screws, 2 mounting brackets, 2 DIN rail adaptors)  
• 1 Omega screwdriver  
• 1 OM-DAQ-USB-2400 Series User Manual (this one, M-5025)  
• 1 OM-DAQ-USB-2400 Series Quickstart Manual (MQS-5025)  
• 2 Type K Thermocouples with Stripped Leads  
2-1  
3
Hardware Setup  
Section 3 - Hardware Setup  
3.1 OM-DAQ-USB-2401 System Components  
127 (5.00)  
M C M L  
H
8
A
M L  
O 7  
H
7
A
M C M L  
H
6
A
M L  
O 5  
H
5
A
O
C
X
E
O 8  
O
C
X
E
O 6  
C
A
C
A
C
A
C
A
A N A L O G I N P U T  
C H A N N E
A N A L O G I N P U T  
C H A N N E L  
1 2 V O U T  
1 2 V O U T  
107  
(4.20)  
!
7 & 8  
5 & 6  
97  
(3.80)  
OM-DAQ-USB-2400 SERIES  
USB DATA ACQUISITION SYSTEM  
CHANNEL 1 & 2  
CHANNEL 3 & 4  
12 V OUT  
!
12 V OUT  
ANALOG INPUT  
ANALOG INPUT  
A
1
H
A
1
L
C
O
M
A
2
H
A
2
L
C
O
M
E
X
C
C
O
M
A
3
H
A
3
L
C
O
M
A
4
H
A
4
L
C
O
M
E
X
C
C
O
M
-
+
38  
(1.5)  
EXT  
PWR  
9V  
1.0 A  
PWR  
IND  
USB  
STATUS  
POWER SUPPLY  
POWER LED  
STATUS LED  
USB PORT  
DIMENSIONS mm (in)  
Figure 3-1. OM-DAQ-USB-2401 Hardware and Dimensions  
NOTE:  
The external power port is used in applications which have  
USB hubs that are not self-powered. See the next section for  
more information.  
NOTE:  
The 'STATUS LED' blinking at Scan Rate means an acquisition  
is in progress. At a fast scan rate, the 'STATUS LED' appears  
solid. The 'POWER LED' in solid means the unit has power  
3.2 USB Hubs and Power Adaptors  
With the use of USB hubs you can connect up to 10 DAQ units to one PC. USB  
hubs can be of the self-powered type, or of the USB-powered type. Both types of  
hubs are available from a variety of vendors.  
3-1  
3
Hardware Setup  
3.2.1 USB-powered Hubs  
These hubs draw all power from the host USB connector’s power pins. The  
power is used for hub internal functions and for the hub’s ports. Each port of a  
USB-powered hub must be capable of supplying at least 100 mA.  
3.2.2 Self-powered Hubs  
These hubs draw power from a source other than the host USB connector, with  
exception that they may draw up to 100 mA from their upstream connection for  
hub internal functions. The external power is used for hub internal functions and  
for the hub’s ports. Each port of a self-powered hub must be capable of  
supplying 500 mA.  
3.2.3 Power Adaptors  
Power adaptors, also referred to as auxiliary power packs, are required for some  
self-powered hubs, and for DAQ modules that are powered from USB-powered  
hubs. In addition, DAQ units will require the use of a power adaptor when used  
with certain laptops.  
NOTE:  
When using a power adaptor with your DAQ system, be sure  
to supply power (from the adaptor to the DAQ unit) before  
connecting the USB cable. This allows the DAQ to inform the  
host computer (upon connection of the USB cable) that the  
unit requires minimal power from the computer.  
Power adaptors for use with the DAQ have a current limit of 500 mA (min.) and  
a voltage range of +7.5 to +12 Volts DC. These specifications are provided on the  
end-face of the DAQ unit.  
If the computer does not recognize the DAQ unit, make sure the computer’s USB  
port is properly enabled and is in good working order. If the computer still fails  
to recognize the DAQ unit, the use of an external 9v DC power adaptor may be  
required.  
NOTE:  
The use of certain notebook computers may require the use  
of a power adaptor with your DAQ unit.  
3-2  
3
Hardware Setup  
3.3 Connecting Your OM-DAQ-USB-2401 Acquisition System to PC  
NOTE:  
When using a power adaptor with your DAQ system, be sure  
to supply power (from the adaptor to the DAQ) before  
connecting the USB cable. This allows your DAQ to inform  
the host computer (upon connection of the USB cable) that the  
unit requires minimal power from the computer.  
Use an approved high-speed USB cable to connect the DAQ system to one of the  
host computer’s USB ports. There is no need for an additional power source in  
this setup since the power pins (of the PC’s USB connection) supply 500 mA at 4  
to 5.25 V. Additional setup examples are described in this chapter, some of which  
involve USB hubs and/or power adaptors.  
POWER  
ADAPTER  
(OPTIONAL)  
USB  
Figure 3-2. Connection of an OM-DAQ-USB-2401 System to a Computer USB Port  
NOTE:  
Certain notebook computers require the use of a power  
adaptor with your DAQ unit.  
3-3  
3
Hardware Setup  
3.4 Connecting Various Hardware Setups  
ON-DAQ-USB-2401 data acquisition systems range from simple to complex. One  
example of a simple system is that of one OM-DAQ-USB-2401 connected to a  
PC’s USB connector. A much more complex system is one that contains 10 OM-  
DAQ-USB-2401 units and a combination of USB-powered and self-powered  
hubs. Despite the wide range of possibilities in between, use of the following  
examples should enable you to properly connect your system.  
NOTE:  
In the examples that follow, the USB hubs have four external  
ports (downstream ports). The USB hubs used in your system  
may have more. Connections can be adjusted accordingly.  
3.4.1 Direct Connection to Computer USB Port(s)  
In this example, one DAQ unit is connected by cable to one of the computer’s  
USB ports. The number of USB connectors may vary from PC to PC. When you  
connect DAQ units directly to a USB connector in this manner, no additional  
power source is required since the computer’s USB connector power pins supply  
the DAQ with adequate power (500 mA at 4.75 to 5.25 V).  
POWER  
ADAPTER  
(OPTIONAL)  
USB  
Figure 3-3. Direct Connection to a Computer USB Port(s)  
3-4  
3
Hardware Setup  
3.4.2 Connection to USB-powered Hub  
In this example, two OM-DAQ-USB-2401s are connected by cable to individual  
ports of a single USB-powered hub. Since the hub receives all its power from the  
computer’s USB, the hub cannot supply adequate power to the DAQ units.  
Because of this aspect of insufficient power, each unit is connected to its own  
power adaptor.  
POWER  
ADAPTER  
POWER  
ADAPTER  
USB  
HUB  
USB  
Figure 3-4. Connection to USB-Powered Hub  
NOTE:  
The power adaptors used must be capable of supplying at  
least 500 mA and have a voltage rating of +7.5 to +12 Vdc.  
3.4.3 Connections to Self-Powered and USB-powered Hubs  
This example illustrates a system that makes use of four DAQ units and two  
different style USB hubs. Two DAQs are connected by cable to individual ports  
of a self-powered USB hub. In addition, the self-powered hub is connected to a  
downstream USB-powered hub, which is also connected to two DAQ units.  
Notice that the DAQ units connected to the self-powered hub have no adaptors  
connected to them. This is because the hub receives external power (in addition  
to the PC supplied USB power), which is capable of supporting the downstream  
devices connected directly to it. In comparison, the three OM-DAQ-USB-2401s  
connected to the USB-powered hub each require their own power adaptor. As in  
example 3.4.2, the power adaptors used must be capable of supplying at least  
500 mA and have a voltage rating of +7.5 to +12 Vdc.  
3-5  
3
Hardware Setup  
POWER  
ADAPTER  
POWER  
ADAPTER  
USB  
HUB  
USB  
POWER  
ADAPTER  
USB  
HUB  
Figure 3-5. Connections to Self-Powered Hub and to USB-Powered Hub  
NOTE:  
USB port locations vary from PC to PC.  
3.5 OM-DAQ-USB-2401 Mounting  
The OM-DAQ-USB-2401 mounting kit contains the following accessories:  
• 2 mounting brackets (to be used for either wall or DIN rail)  
• 4 long screws (to be used for either wall or DIN rail)  
• 4 short screws (to be used for the DIN rail)  
• 2 DIN rail plastic adaptors  
3-6  
3
Hardware Setup  
3.5.1 OM-DAQ-USB-2401 Wall Mounting  
In order to mount the OM-DAQ-USB-2401 to the wall, you will need to:  
1. Remove the rear screws from the OM-DAQ-USB-2401.  
2. Align the mounting brackets with the rear holes on the DAQ unit (make sure  
that the ears are facing outwards, like in the figure below).  
3. Use the longer screws to reconnect the mounting brackets to the body of the  
enclosure, in the same direction as the screws that were removed from the  
DAQ unit.  
4. Drill 2 holes for #4 screws, 5.50" apart. Tip: Hold the DAQ unit with mounting  
brackets attached, up against the wall; use the holes as a guide to mark where  
you will drill holes in the wall.  
5. Install the two screws into the wall.  
6. Mount the DAQ unit on the wall by passing the screw head through the larger  
mounting bracket holes and then sliding it down.  
Figure 3-6. OM-DAQ-USB-2401 Wall Mounting  
146 (6.00)  
133 (5.50)  
MOUNTING HOLE DISTANCE  
M C M L  
H
8
A
M L  
O 7  
H
7
A
M C M L  
H
6
A
M L  
O 5  
H
5
A
O
C
X
E
O 8  
O
C
X
E
O 6  
C
A
C
A
C
A
C
A
A N A L O G I N P U T  
C H A N N E L  
A N A L O G I N P U T  
C H A N N E L  
1 2 V O U T  
1 2 V O U T  
!
7 & 8  
5 & 6  
Ø3  
2 PLACES  
(0.125)  
OM-DAQ-USB-2400 SERIES  
107  
(4.20)  
USB DATA ACQUISITION SYSTEM  
Ø6  
2 PLACES  
(0.25)  
CHANNEL 1 & 2  
CHANNEL 3 & 4  
12 V OUT  
!
12 V OUT  
ANALOG INPUT  
ANALOG INPUT  
A
1
H
A
1
L
C
O
M
A
2
H
A
2
L
C
O
M
E
X
C
C
O
M
A
3
H
A
3
L
C
O
M
A
4
H
A
4
L
C
O
M
E
X
C
C
O
M
Figure 3-7. OM-DAQ-USB-2401 Wall Mounting Dimensions  
3-7  
3
Hardware Setup  
3.5.2 OM-DAQ-USB-2401 DIN Rail Mounting  
In order to mount the OM-DAQ-USB-2401 to a DIN-Rail, you will need to:  
1. Attach the DIN-rail plastic adaptors to the mounting brackets, using the four  
shorter screws.  
2. Remove the rear screws from the OM-DAQ-USB-2401.  
3. Align the mounting brackets with the rear holes on the DAQ unit (make sure  
that the ears are facing inwards, like in the figure below).  
4. Use the longer screws to reconnect the mounting brackets to the body of the  
enclosure, in the same direction as the screws that were removed from the  
DAQ unit.  
5. Attach the DAQ unit to the DIN rail, using the plastic adaptors attached to the  
mounting brackets.  
Figure 3-7. OM-DAQ-USB-2401 Rail Mounting  
3-8  
4
Software  
Section 4 - Software  
4.1 Getting Started  
The following program files are included on the Omega DAQ Central User  
Software CD supplied with your data acquisition unit. These files can also be  
downloaded from the omega.com website should you misplace your CD.  
• DAQ Central data logging software installer  
• Omega.DAQ .NET API  
NOTE:  
When using Omega DAQ modules to acquire data, computer  
energy save modes can cause false data readings. Prior to  
using Omega DAQ modules, ensure your computers energy  
save mode is disabled. If needed, consult your PC users  
manual to disable energy save (power suspension) modes.  
4.2 Software Installation  
4.2.1 System Requirements  
Minimum Requirements:  
• Processor: 800 MHz  
• Hard Drive Space: 300 megabytes  
• Ram: 512 megabytes or higher  
• 1 Available USB Port  
• CD-ROM Drive or Internet connection  
• Windows XP Service Pack 3 (32-bit), Windows Vista (32-bit), Windows 7 (32-  
bit), or Windows 7 (64-bit)  
Recommended Requirements:  
• Processor: 1.5 GHz  
• Hard Drive Space: 300 megabytes  
• Ram: 1.5 gigabytes  
• 1 Available USB Port  
• CD-ROM Drive or Internet connection  
• Windows XP Service Pack 3 (32-bit), Windows Vista (32-bit), Windows 7 (32-  
bit), or Windows 7 (64-bit)  
4-1  
4
Software  
4.2.2 Software Setup  
Insert the DAQ Central Software CD that was included with your DAQ unit into  
the CD-ROM drive on your PC. Your system should begin the installation  
process automatically. If the software installation does not start automatically:  
open: "My Computer", double-click your CD-ROM drive, and run "setup.exe".  
Figure 4-1. Welcome Screen  
This welcome screen should be visible on your computer screen. To continue  
with installing the program click the “Next >” button.  
From this screen you select the folder were you want the program files installed  
on your PC. The default setting will install the software under your “Program”  
folders in a new folder named “Omega” To continue with installing the program  
click the “Next >” button.  
Figure 4-2. Select Install Screen  
4-2  
4
Software  
Figure 4-3. Confirm Installation Screen  
The setup wizard now has all the information to complete the installation of the  
software on your PC. To continue with installing the program click the “Next >”  
button.  
During the install process, a Windows Command Prompt may pop up. No  
interaction is required in this window; it is automatically installing or updating  
the device’s USB drivers. Please let this process complete before closing the  
software installation wizard.  
Congratulations! You have just successfully installed the DAQ Central Program  
on your PC. To end installing the program and close the setup wizard click the  
“Close” Button.  
4-3  
4
Software  
4.3 DAQ Central Software Operation  
Below is the basic flow of operation when using the software:  
1. Highlight desired device.  
2. Open Device Configuration window.  
3. Set channels, sample rate, triggering, and data output file.  
4. Click “OK” to confirm Device Configuration.  
5. Open any desired Data Displays.  
6. Click the Start button.  
Details on the various windows and options can be found in the rest of this  
chapter.  
Figure 4-4. Main Control Window  
4-4  
4
Software  
TOOLBAR BUTTONS  
Start Button  
Initiates data acquisition for the  
highlighted device. Even if a Start  
Trigger is set, this must be pressed  
to Arm the device.  
Stop Button  
Stops data acquisition for the  
highlighted device. If a Stop  
Trigger is set, this will cancel the  
trigger.  
Device  
Configuration  
Icon  
Accesses the Device Configuration  
window for the selected device.  
The Device Configuration window  
is used to set up all device options,  
including channels, sample rate,  
triggers, and data output file.  
Gauges Display Opens a Gauges window.  
Button  
Displays measurements in analog  
meter format.  
Bars Display  
Button  
Opens a Bars window. Displays  
measurements in a bar graph  
format.  
Digital Display  
Button  
Opens a Digital meter window.  
Displays measurements in a digital  
meter format.  
Table Display  
Button  
Opens a Table window. Displays  
measurements in a table format.  
Waveforms  
Display Button  
Opens a Waveform window.  
Displays the measurements in a  
scrolling chart.  
Device Display  
List  
Displays:  
- The available connected DAQ  
devices - DAQ2401DEMO –  
Demo/simulation device for testing  
software features without  
hardware.- LEDs – Will flash green  
during the data acquisition.  
Table 4-1. Toolbar Buttons  
4-5  
4
Software  
4.3.1 Main Control Window Pull-Down Operations  
Below are the operations that are available in the Pull-Down Menus of the DAQ  
Central Main Control Window.  
Figure 4-5. File Menu  
Open Configuration  
Opens Device Configuration settings from a saved file  
Save Configuration As Saves current Device Configuration settings to a new file name  
These settings files contain all of the data that is found in the Device  
Configuration window.  
Figure 4-6. Device Menu  
Detect Devices  
Configuration  
Detects available OM-DAQ-USB-2401 devices which are currently  
connected to the computer through the USB.  
Same function as Configuration button  
Accesses the device  
.
Configuration window to configure ‘Channels’ settings, ‘Triggers’  
settings, and ‘Data’ settings.  
Start Acquisition Same function as Start button  
. Initiates data acquisition for  
.
the highlighted device. Even if a Start Trigger is set, this must be  
selected to Arm the device.  
Stop Acquisition Same function as Stop button  
. Stops data acquisition for the  
highlighted device. If a Stop Trigger is set, this will cancel the trigger.  
4-6  
4
Software  
Figure 4-7. Tools Menu  
Export Data File – To convert data file from binary file to .csv file  
Fgure 4-8. Help Menu  
Check for Updates To check for the latest Omega DAQ Central software update  
information through the Internet connection.  
About  
To view software version.  
4-7  
4
Software  
Section 4.3.2 Device Configuration  
The Device Configuration window contains the Channel, Triggers, and Data  
tabs. Details for these tabs are below. When finished making changes, press  
“OK” to confirm the changes or “Cancel” to abandon the changes.  
Figure 4-9. Configuration Mode - Channels Tab Screen  
Channels Tab  
The Channels tab contains a spreadsheet designed for setting up channels for  
data acquisition. Depending on the column, you can make changes to the  
information contained in a cell by placing the mouse cursor in the cell.  
All columns except “Channel” are editable. Those with drop-down arrows are  
limited to the choices in the drop-down list.  
You can use your PC’s keypad arrow and tab keys to select new “active cells” in  
the spreadsheet.  
4-8  
4
Software  
Configuration Parameter Settings  
Channel The input channels name (not editable)  
AN1D = analog channel 1, differential input  
AN1H = analog channel 1, single-ended high input  
AN1L = analog channel 1, single-ended low input  
Etc.  
Nickname You can specify the Nickname (e.g., Oven 1, Monitor  
Room)  
Units  
You can specify the Engineering Unit. When switching  
between voltage and thermocouple ranges, this will  
automatically switch to the default units for that range.  
However, any units label may be typed in here.  
On/Off  
To turn on or turn off the selected input channel  
Single/Diff You can select input channel as Single-ended or  
Differential input.  
Range  
You can select the voltage input range or thermocouple  
type for each input channel. Thermocouples must  
always be used with Differential input.  
Scale and You can rescale the input signal. This scaled/offset  
Offset  
value will appear in all data displays, and in any saved  
data files.  
Sampling can be set in terms of “Scan Rate” or “Scan Time”.  
Scan Rate = 1/(Scan Time).  
For example, if you are acquiring 4 voltage inputs at 250 Hz, you will receive 250  
complete scans per second. Meaning each channel will send 250 pieces of data  
every second, for a total of (4 x 250) = 1000 pieces of data.  
The maximum scan rate is affected by the number of channels turned on.  
Selecting thermocouple input also lowers the maximum scan rate.  
Figure 4-10. Scan Options  
Figure 4-11. Units Options  
Figure 4-12. Automatic Calibration  
4-9  
4
Software  
Figure 4-13. Trigger Options  
The Triggers tab is used to set up triggers for starting and stopping data  
acquisition. The Start Trigger and the Stop Trigger each have a different set of  
available trigger types:  
Figure 4-14. "Start Trigger" Options  
Figure 4-15. "Stop Trigger" Options  
Trigger Types  
Manual  
Acquisition is manually started/stopped.  
Level  
Set trigger to stop/start acquistition when the input level is either above or below a certain threshold.  
Date/Time Set trigger to stop/start acquistition at a certain date and time, as per the PCs clock.  
Time Span Set trigger to stop acquistition after a certain elapsed time.  
Scan Count Set trigger to stop acquistition after a certain number of scans have been acquired.  
4-10  
4
Software  
NOTE:  
For all Start Trigger types, the device must be armed by pushing the Start Button on the main window.  
Data storage and display will not begin until the device triggers.  
Data Tab  
Figure 4-16. Data Tab  
To save data to a file on the PC, select the Data tab and choose a filename. This data  
can be saved in Binary or Delimited (.csv) format.  
• Binary format is faster, and is recommended for slower PCs. Binary format also  
creates files that are about 50% smaller than Delimited files. Binary files can be  
converted to “.csv” format later by selecting Tools --> Export Data File.  
• Delimited format can use either commas or tabs as the delimiter, and the time can  
be saved as Elapsed Time or as a Date/Time stamp.  
There are three “Save Behavior” options:  
FIG 4-17. Data ‘Save Behavior’ Options  
Auto-overwrite  
If a file with the same name already exists, it will be automatically overwritten when  
the Start Acquisition button is pressed.  
Prompt for  
overwrite  
If a file with the same name already exists, a prompt  
will appear when the Start Acquisition button is pressed.  
Add creation  
time to  
filename  
A timestamp is added to the filename, in the format  
YYYY-MM-DD-hhmm-ss. For example, MyFile.bin  
becomes MyFile-2011-1031-1121-03.bin. The timestamp will be automatically added  
every time the Start Acquisition button is pressed, so multiple acquisitions can be easily  
stored to the same folder without revisiting the Device Configuration window.  
Data Displays  
From the Main Window, you can choose to view live data from any combination of Gauges, Bars,  
Digital Displays, Tables, and Waveforms. These displays can be opened by clicking the appropriate  
button in the Main Window.  
4-11  
4
Software  
Figure 4-18. Main Window Screen  
Figure 4-19. Gauges Screen  
Figure 4-20. Bars Screen  
Figure 4-21. Digital Screen  
Figure 4-22. Table Screen  
4-12  
4
Software  
Figure 4-23. Waveform Screen  
Each Data Display can be customized by clicking the Options Icon  
.
Each Data Display's Options screen lets you choose which device channels to  
display in that window. Channels can be added or removed with the Add  
Display  
and Delete Display  
buttons. Confirm changes by clicking the  
OK button. Start Acquistion by clicking the Play button on the Main Window.  
Acquistion can be stopped with the Stop button.  
Figure 4-24. Sample Options Screen  
4-13  
4
Software  
4.4 Power Line Rejection  
The OM-DAQ-USB-2401 Series can take readings while making use of 50/60Hz  
line cycle rejection. While in the line cycle rejection mode, the maximum sample  
rate is as follows:  
• 50 Hz rejection: Scan Rate (SR) Range: 0.153Hz < SR <= 1Hz  
• 60 Hz rejection: Scan Rate (SR) Range: 1Hz < SR <= 1.667Hz  
OMB-DAQ-USB-2401 Series Block Diagram  
Power LED  
Isolation  
Barrier  
Power Supply  
Isolated  
Power Supply  
Power Jack  
Local Power  
USB  
Jack  
USB  
Controller  
MCU  
32Bit  
Activity LED  
Digital Isolators  
CJC  
4
Analog Inputs  
Gain =  
x2, 1/8  
+12v  
+12v output  
Excitation voltage  
SE/  
DI  
MUX  
PGIA  
CJC  
4
Analog Inputs  
Gain =  
x1,2,4,8,16  
32,64  
+12v output  
Excitation voltage  
24–Bit  
A/D  
PGA  
MUX  
1MΩ  
1MΩ  
CJC  
4
Analog Inputs  
Gain =  
x2, 1/8  
+12v  
+12v output  
Excitation voltage  
SE/  
DI  
PGIA  
MUX  
CJC  
4
Analog Inputs  
+12v output  
Excitation voltage  
1MΩ  
1MΩ  
Figure 4-25. OM-DAQ-USB-2400 Block Diagram  
4.5 Optical Isolation  
Your DAQ unit is optically isolated from its host PC by up to 500 Vdc. This  
means that an inadvertent application of such voltage to the DAQ unit will not  
affect the PC. In addition to isolation, the device maintains all sensitive  
acquisition-related circuitry external to the PC. This physical isolation of  
circuitry from the PC results in less noise and more accurate measurements.  
4.6 A/D Conversion  
OM-DAQ-USB-2401 uses an extremely low-noise, 24-bit analog-to-digital (A/D)  
converters. They provide complete high-resolution measurement solutions for  
the most demanding applications.  
4-14  
4
Software  
The converter is comprised of a 4th-order, delta-sigma (Δ∑) analog-to-digital  
(A/D) converter modulator followed by a programmable digital filter. A flexible  
input multiplexer handles differential or single-ended signals; the programmable  
filter allows the user to optimize between a resolution of up to 23 bits noise-free.  
Examples of scan sequences, with various channel resolutions and calibration  
arrangements, appear in Table 5-1 on page 5-2.  
4.7 Input Ranges  
You can individually select the input range for each channel. For example, one  
channel could be used for volts and another for temperature. Omega’s OM-  
DAQ-USB-2401 automatically assigns the appropriate units depending on two  
factors:  
• The selected range and  
• Measurement unit preferences  
Measurement unit preferences can be modified from the Omega DAQ Central  
‘Channels’ tag of the ‘Configuration’ window. Chapter 4 provides more detailed  
information.  
NOTE:  
The maximum voltage input range (full scale) is –10 to +10  
Vdc. The lowest programmable voltage input range is -30mV  
to +30 mV. A complete list of the OM-DAQ-USB-2401s  
programmable ranges appears on page 8-1.  
Each analog input channel has the following user-specified measurement  
parameters:  
• Signal type: volts, or thermocouple type J, K, T, E, R, S, B, or N  
• Full scale voltage: from -10 to + 10 Vdc; with programmable ranges as  
indicated on page 8-1.  
• Resolution/sample period: from 22.5 bit RMS at 0.153 samples/sec, to 17 bit  
RMS at 50 samples/sec or greater  
NOTE:  
These rates were obtained with a 16-channel scan.  
4.8 Analog Input Configuration  
The OM-DAQ-USB-2401 includes 16 analog signal inputs which may be used as  
16 single-ended inputs, 8 differential inputs, or as a combination of single-ended  
and differential inputs with up to 16 connections.  
Single-ended inputs are used with signals that share the same common low  
(COM), such as multiple batteries which have their negative sides connected in  
common. Differential inputs are required when signals do not share the same  
common low, such as in the typical use of thermocouples.  
4-15  
4
Software  
NOTE:  
In DAQ applications, thermocouples should not be connected  
single-ended. Doing so can result in noise and false readings.  
This is especially true when acquiring other high-amplitude  
signals in conjunction with thermocouple signals that are  
connected single-ended.  
The DAQ units include built-in cold-junction compensation (CJC), which is  
automatically invoked when you select TC measurements. The OM-DAQ-USB-  
2401 automatically converts acquired voltage readings into compensated-  
linearized temperature readings. This DAQ system can make thermocouple and  
volts measurements concurrently.  
OM-DAQ-USB-2401  
TERMINAL BLOCK  
(PARTIAL)  
A
1
H
A
1
L
C
O
M
A
2
H
A
2
L
V1  
V2  
COM  
V3+  
V3  
V
3-  
Figure 4-26. OM-DAQ-USB-2401 Single-Ended and Differential Connections to  
Analog Input Channels Diagram  
Figure Notes: *V1 and V2 are single-ended inputs.  
*V3 is a thermocouple and is a differential input.  
4.9 Measurement Duration, Scan Time, and Resolution  
In relation to sampling analog input, the terms measurement duration, sample  
rate, and resolution have the following meanings:  
Measurement duration (per channel) – the amount of time used for sampling a  
channel’s input signal. The measurement durations for the OM-DAQ-USB-2401's  
analog channels range from very slow (400 milliseconds for one measurement  
duration) to very fast (1 millisecond for one measurement duration).  
Scan Time (per scan) - The amount of time used for scanning (sampling) all  
selected channels’ input signal. For the OM-DAQ-USB-2401, the scan rates range  
from very slow (1 hour per scan) to very fast (1 milliseconds per scan).  
4-16  
4
Software  
Resolution (Effective Number of Bits - ENOB, RMS) – The number of reliable  
data bits that exist for a signal’s measurement. The greater the resolution, the  
more detailed the reading. For example, with an increased resolution, a reading  
of 5.12 V could become 5.11896 V. The DAQ actually provides for 24 bits of data  
information; however, the accuracy of the least significant bits decreases as the  
measurement duration speeds up.  
NOTE:  
When measuring variable input signals (as opposed to  
relatively steady input signals), the variable signals will  
require more samples/sec to obtain a realistic signal  
representation.  
At a scan time greater than 6.5 seconds, the last 1.5 bits are considered unreliable,  
resulting in a resolution of 22.5 bits. At a very fast scan time (1 millisecond), the  
seven least significant bits are unreliable, resulting in 17 bit accuracy. When you  
select the Scan Time you also determine the scan (sample) rate and resolution for  
the applicable channel. For the DAQ’s analog input applications, scan (sample)  
rates range from 0.0002778 samples/sec (1 hr/scan) up to 1000 samples/sec  
(m/scan), and corresponding resolution ranges from 22.5 to 17 bits.  
4.10 Automatic Calibration  
The DAQ unit contains a built-in source for performing automatic self-  
calibrations. These calibrations can be performed between scans periodically  
throughout the measurement process, as indicated in the following figure. Such  
calibration ensures accurate measurements, even in environments that  
experience significant temperature fluctuations. The figure below shows an  
initial calibration followed by the scan/calibrate, scan/calibrate pattern.  
NOTE:  
The continuous calibration feature is built into the DAQ  
system. With Auto-Calibration enabled, you may experience  
slower sample rates especially with multiple chanels turned on.  
INITIAL CALIBRATION  
SCAN  
CALIBRATION  
SCAN PERIOD  
SCAN  
CALIBRATION  
SCAN  
CALIBRATION  
TIME  
0
Figure 4-27. Calibration/Scan Arrangement  
4-17  
4
Software  
4.11 Thermocouple Measurements  
The OM-DAQ-USB-2401 provides effortless thermocouple (TC) measurements.  
The unit includes built-in cold-junction compensation (CJC), which is  
automatically invoked when you select TC measurements. The DAQ unit  
automatically converts acquired voltage readings into compensated-linearized  
temperature readings. A DAQ system can make thermocouple and volts  
measurements concurrently.  
TIP: If you experience noise from thermocouple wires, you can connect a 100K  
resistor from AxL to the COM to elimate some noise. Please reference the  
following diagram:  
CONNECT A 100K RESISTER  
TO ELIMATE SOME  
THERMOCOUPLE INPUT NOISE  
A
1
H
A
1
L
C
O
M
A
2
H
A
2
L
100K  
V3+  
V3  
V
3-  
Figure 4-28. Thermocouple Wiring  
4-18  
5
Signal Management  
Section 5 – Signal Management  
5.1 Channel Control and Expansion  
In the OM-DAQ-USB-2401 system, the quantities and types of DAQ units used  
determine the system’s channel capacity. Up to 10 OM-DAQ-USB-2401s can be  
used with one host PC. See the Specifications for more detailed channel  
information.  
5.2 Scan Time and Resolution  
5.2.1 Scan Time  
Scan Time (per scan) - The amount of time used for scanning (sampling) all  
selected channels’ input signal. For the OM-DAQ-USB-2401, the scan rates range  
from very slow (1 hour per scan) to very fast (1 milliseconds per scan).  
5.2.2 Resolution (Effective Number of Bits – ENOB, RMS)  
Resolution (Effective Number of Bits - ENOB, RMS) – The number of reliable  
data bits that exist for a signal’s measurement. The greater the resolution, the  
more detailed the reading, for example, with increased resolution a reading of  
5.12 V could become 5.11896 V. The DAQ actually provides for 24 bits of data  
information; however, the accuracy of the least significant bits becomes less as  
the measurement duration speeds up.  
At a scan time greater than 6.5 seconds, the last 1.5 bits are considered unreliable,  
resulting in a resolution of 22.5 bits. At a very fast scan time ( 1 milliseconds), the  
seven most least significant bits are unreliable, resulting in 17 bit accuracy.  
When you select the Scan Time you also determine the scan(sample) rate and  
resolution for the applicable channel. For the DAQ’s analog input applications,  
scan(sample) rates range from 0.0002778 samples/sec up to 1000 samples/sec  
and corresponding resolution ranges from 22.5 to 17 bits.  
5-1  
5
Signal Management  
2
Scan Rate vs. Resolution  
Resolution  
3,4  
5
Scan Rate (Scans  
(ENOB  
1
Speed Designation  
Scan Time (s)  
per Second - SPS)  
RMS)  
Very Slow, 50 / 60 Hz rejection  
3600s>=ST >= 6.5s  
0.0002778=< SR <=  
0.153  
22.5  
Slow, 50 Hz rejection  
Slow, 60 Hz rejection  
Medium, 60 Hz rejection  
Medium, 50 Hz rejection  
Medium  
6.5s> ST >= 1s  
1s> ST >=0.6s  
0.153 < SR <=1  
1 < SR <= 1.67  
1.67 < SR <= 3.33  
3.33 < SR <= 5  
5 < SR <= 20  
21  
20.9  
20.4  
20.3  
19.2  
18.7  
17  
0.6s> ST >=0.3s  
0.3s> ST >=0.2s  
0.2s> ST >=0.05s  
0.05s> ST >=0.02s  
0.02s> ST >=0.001s  
Fast  
20 < SR <= 50  
50< SR <= 1000  
Very Fast  
Table 5-1. Scan Rate vs. Resolution (ENOB, RMS)  
Notes:  
1. Scan Rate = 1/(Scan Time);  
2. Each channel can have independent measurement duration and resolution.  
3. The scan rates and resolutions shown were obtained with a 16-channel scan and with continuous  
self-calibration.  
4. When measuring TC input, the CJC measurements will be treated as a channel.  
5. ENOB, Effective Number of Bits  
5.3 Under Sampling and Aliasing  
When you select the Scan Time you also determine the scan rate and resolution  
for the applicable channel. As shown in the previous table, scan rates range from  
0.0002778 scans/sec up to 1000 scans/sec and corresponding resolution ranges  
from 22.5 to 17 bit (for the DAQ’s analog input applications).  
When measuring highly variable input signals (as opposed to relatively steady  
input signals), the variable signals will require more scans/sec to obtain a  
realistic signal representation. When too few samples are taken, the term under  
sampling is often used. Under sampling tends to lower the signal’s graphed  
amplitude. This is also known as aliasing. In such cases, the inadequate number  
of samples results in a “flattening” of the signal. Gross under sampling can even  
result in a relatively flat line, even though the actual signal is a waveform.  
For best results the sample rate (scan rate) should be at least 10 times the  
measurement frequency. If possible it is recommended that you set the scan rate  
to 20 times the frequency of the input signal.  
5-2  
5
Signal Management  
The following table provides general advice regarding the selection of Scan  
Time. The concepts are further illustrated by the figure, Examples of Under  
Sampling.  
Scan Time  
(Measurement  
Duration)  
Scan Rate  
(Sample Rate)  
Analog Input Signal  
Resolution  
High  
Steady, or gradual  
change  
Long  
Low  
Highly variable  
(unsteady)  
Short  
High  
Low  
Figure 5-1. Examples of Under Sampling  
The above figure depicts three signals from the same temperature fluctuations.  
Under Sampling (aliasing) is evident in two of the signals. Elaboration follows:  
Signal 1:  
This signal, based on 1 sample(scan) per division, is represented by a heavy solid  
line and sample-points designated by polygon symbols. The signal represented  
is a fairly accurate portrayal of the actual temperature fluctuations.  
Signal 2:  
This signal, based on 1 sample(scan) every 3 divisions, is represented by a heavy  
dotted line and sample-points designated by squares. Under sampling has  
resulted in a distortion, in effect, a lower amplitude than that exhibited by the  
first signal, even though each measured point is accurate.  
Signal 3:  
This signal, based on 1 sample(scan) every 4 divisions, is represented by a  
dotted/dashed line and sample-points designated by plus signs (+). Fewer  
samples (a greater degree of under sampling) has resulted in a further distortion  
and lowering of signal amplitude.  
From these examples it should be realized that more samples(scans) will result in  
a more accurate representation of the actual signal, and that under sampling will  
tend to lower the amplitude of the signal, exhibiting a trend toward a “flat line”  
state.  
As stated earlier, the more variability a signal has, the more samples that are  
needed to accurately portray it.  
5-3  
5
Signal Management  
5.4 Triggering  
Triggering controls an acquisition cycle. Once the system is armed, a trigger is  
required to collect the data.  
The user must determine the triggering requirement based on the nature of the  
measurement and the amount of data needed to satisfy the system’s purpose.  
The trigger source can be a software command or an analog input channel on  
reaching a specified voltage level can be used to trigger the system.  
IMMEDIATE  
TYPE OF  
START TIME  
TRIGGER SOURCE  
CHANNEL VALUE - RISING  
- FALLING  
TRIGGER EVENT  
POST-TRIGGER SCAN COUNT  
SCAN GROUP  
TIME  
SCAN PERIOD  
Figure 5-2. Trigger Source Diagram  
5.5 Signal Modes  
OM-DAQ-USB-2401 units operate in one of two modes, (1) single-ended mode,  
or (2) differential mode. These terms (single-ended mode and differential mode)  
apply to their use in this manual. In other sources these terms may be used in a  
different manner.  
Choosing between differential and single-ended inputs is made by software  
command. The following text briefly describes the two signal modes.  
Single-ended mode refers to a mode, or circuit set-up, in which a voltage is  
measured between one signal line and common ground voltage (Vcm). The  
measured voltage may be shared with other channels. The advantage of a single-  
ended non-differential mode [over differential mode] is that it provides for a  
higher channel count, for example: 16 channels instead of 8.  
NOTE:  
In DAQ applications, thermocouples should not be connected  
single-ended. Doing so can result in noise and false readings.  
This is especially true when acquiring other high-amplitude  
signals in conjunction with thermocouple signals that are  
connected single-ended.  
5-4  
5
Signal Management  
Differential mode refers to a mode, or circuit set-up, in which a voltage is  
measured between two signal lines. The measured differential voltage is used for  
a single channel. An advantage of using differential inputs is that they reduce  
signal errors and the induction of noise resulting from ground current. The  
following illustration is an example of how noise is reduced, or canceled-out,  
when using the differential  
Figure 5-3. Noise Reduction in Differential Mode  
In the schematic, voltage signal S2 is subtracted from signal S1, resulting in the  
output signal shown. The noise spikes (having the same polarity, phase, and  
magnitude in each input signal) cancel each other out. This results in a clean  
differential signal (S1 - S2).  
Floating-differential measurements are made when low-level signals must be  
measured in the presence of high levels of common-mode noise (e.g., a non-  
grounded thermocouple). When the signal source has no direct connection to the  
system analog common, one must be provided. In Omega Daq the connection to  
analog common is provided in the circuitry with both the channel high and  
channel low connected to analog common. Both of these connections to common  
are made through 1 MΩ resistors. No additional connections of channel high and  
low to common should be made.  
Figure 5-4. Example of Floating Differential Circuit  
NOTE:  
Differential signal hookups do not provide isolation or any  
kind of circuit protection.  
5-5  
5
Signal Management  
Resolution: An analog-to-digital converter (ADC) converts an analog voltage to  
a digital number. The digital number represents the input voltage in discrete  
steps with finite resolution. ADC resolution is determined by the number of bits  
that represent the digital number. An n-bit ADC has a resolution of 1 part in 2n.  
Thus, 12 and 16 bit resolutions are as follows:  
• 12-bit resolution: 1 part in 4096 (212), corresponding to 2.44 mV in a 10 V  
range.  
• 16-bit resolution: 1 part in 65,536 (216), corresponding to 0.153 mV in a 10 V  
range.  
5.6 System Noise  
Laboratory and industrial environments often have multiple sources of electrical  
noise. An AC power line is a source of 50/60 Hz noise. Heavy equipment (air  
conditioners, elevators, pumps, etc.) can be a source of noise, particularly when  
turned on and off. Local radio stations are a source of high-frequency noise, and  
computers and other electronic equipment can create noise in a multitude of  
frequency ranges. Thus, an absolute noise-free environment for data acquisition  
is not realistic. Fortunately, noise-reduction techniques such as averaging,  
filtering, differential voltage measurement, and shielding are available to reduce  
noise to an acceptable level.  
5.7 Averaging  
Certain acquisition programs apply averaging after several samples have been  
collected. Depending on the nature of the noise, averaging can reduce noise by  
the square root of the number of averaged samples. Although averaging can be  
effective, it suffers from several drawbacks. Noise in measurements only  
decreases as the square root of the number of measurements—reducing RMS  
noise significantly may require many samples. Thus, averaging is suited to low-  
speed applications that can provide many samples.  
NOTE:  
Only random noise is reduced or eliminated by averaging.  
Averaging does not reduce or eliminate periodic signals.  
5.8 Analog Filtering  
A filter is an analog circuit element that attenuates an incoming signal according  
to its frequency. A lowpass filter attenuates frequencies above the cutoff  
frequency. Conversely, a high-pass filter attenuates frequencies below the cutoff.  
As frequency increases beyond the cutoff point, the attenuation of a singlepole,  
low-pass filter increases slowly. Multi-pole filters provide greater attenuation  
beyond the cutoff frequency but may introduce phase (time delay) problems that  
could affect some applications.  
5-6  
5
Signal Management  
5.9 Input and Source Impedance  
As illustrated below, the input impedance (Ri) of an analog-to-digital converter  
combines with the transducer’s source impedance (Rs) forming a voltage  
divider. This divider distorts the voltage being read at the analog-to-digital  
converter. The actual voltage read is represented by the equation:  
VADC = VT x Ri / (Rs + Ri)  
The input impedance (Ri) of most ADCs is at least 1 MΩ; low source impedance  
(Rs) usually presents no problem. Some transducers, such as piezoelectric types,  
have high source impedance, and should therefore be used with a charge-  
sensitive amplifier of low output impedance. As described in the following  
paragraphs, multiplexing can greatly reduce the effective input impedance of an  
analog-to-digital converter.  
Figure 5-5. Analog to Digital Converter  
5.10 Crosstalk  
Crosstalk is a type of noise related to source impedance and capacitance, in  
which signals from one channel leak into an adjacent channel, resulting in  
interference or signal distortion. The impact of source impedance and stray  
capacitance can be estimated by using the following equation.  
T = RC  
Where T is the time constant, R is the source impedance, and C is the stray  
capacitance.  
High source (transducer) impedance can be a problem in multiplexed A/D  
systems. When using more than 1 channel, the channel input signals are  
multiplexed into the A/D. The multiplexer samples each signal and then  
switches to the next input signal. A high-impedance input interacts with the  
multiplexer’s stray capacitance and causes crosstalk and inaccuracies in the A/D  
sample.  
A solution to high source impedance in relation to multiplexers involves the use  
of buffers. The term buffer has several meanings; but in this case, buffer refers to  
an operational amplifier having high input impedance but very low output  
impedance. Placing such a buffer on each channel (between the transducer and  
the multiplexer) prevents the multiplexer’s stray capacitance from combining  
with the high input impedance. This use of a buffer also stops transient signals  
from propagating backwards from the multiplexer to the transducer.  
5-7  
6
CE Conformity  
Section 6 - CE Conformity  
6.1 OM-DAQ-USB-2401 Design for CE Conformity  
The OM-DAQ-USB-2401 has been designed to meet requirements as outlined in  
European Community EMC Directive EN61326. Three ferrite cores are included  
in the OM-DAQ-USB-2401 package. When powering the unit using the adapter,  
the ferrite cores must be attached to both the USB cable and the power adapter to  
meet EFT (Electrical Fast Transient) specifications. Refer to Figure 6-1.  
POWER  
ADAPTER  
USB  
(MIDPOINT OF CABLES)  
Figure 6-1. OM-DAQ-USB-2401 with Ferrite Cores Installed  
To maintain CE certification and install the ferrite cores, perform the following:  
1. Using a small screw driver or other suitable instrument, unsnap the ferrite  
core to open it (as shown in Figure 6-2).  
2. Install one ferrite core on each far end of the USB cable, and one ferrite core in  
the middle of the cable (as shown in Figure 6-1).  
3. Install one ferrite core on each far end of the power adapter, and one ferrite  
core in the middle of the cable (as shown in Figure 6-1).  
6-1  
6
CE Conformity  
UNSNAP  
FERRITE  
CORES  
HERE  
Figure 6-2 Ferrite Core  
The ferrite cores may be removed from the DAQ unit; however, in order to  
conform to CE standards, all of the ferrite cores must be attached to the USB  
cable and power adapter.  
NOTE:  
If the user is not using the power adapter but relying on  
power from the host USB connection, no ferrites are required.  
6-2  
7
Troubleshooting  
Section 7 – Troubleshooting  
7.1 Basic Checklist  
Power – Check USB cable connection. The green “PWR IND” LED should be lit  
on the DAQ device. If using a powered USB hub, make sure its power cable is  
plugged in.  
Signal – Check signal lines and connections. Connectors must be free of  
corrosion. Signal lines should be undamaged and free of sharp bends and twists.  
Signal paths should avoid potential sources of noise (high voltage and  
electromagnetic interference).  
Device Selection – Make sure the serial number selected in the software matches  
the serial number of your device. Each data display will show the serial number  
(or Nickname, if assigned) of the device it is currently displaying.  
“DAQ2401DEMO” is a software demo device which outputs a sine wave.  
Software Updates – A software update may be available. In DAQ Central,  
choose “Check For Updates” from the Help Menu on an Internet-connected PC.  
7.2 Symptoms and Solutions  
USB drivers are not found or hardware installation fails. This can happen if the  
device is plugged into the computer’s USB port before software installation is  
complete. The program “C:\Program Files\Omega\DAQ Central\USB  
Drivers\CDM20814_Setup.exe” can be run to reinstall the USB drivers. All non-  
critical USB devices should be unplugged while this program runs, including all  
DAQ devices.  
The “20814” portion of this filename may be different, and the “DAQ  
Central\USB Drivers” directory may be in a different location if an alternate  
directory was chosen during installation.  
Data does not appear or data does not make sense. See “Device Selection”  
under Basic Checklist.  
Stop the acquisition and check the Triggers tab of the Device Configuration  
window. A trigger may be set to start acquisition at a future date. Try setting  
both Start and Stop triggers to manual and restarting acquisition.  
The Device Configuration screen is grayed out. A data acquisition may be  
running. A device’s configuration can not be edited during acquisition.  
The waveform scale is too small/large The vertical scale can be adjusted by  
clicking the max and min values on the y-axis. The horizontal scale can be adjusted  
from the Waveform Options screen, which is accessed by clicking the Options button  
at the top of the Waveform window. The “Scans to graph” setting determines how  
many data points (scans) will appear along the x-axis. A higher number shows a  
greater span of time, but uses more of the PC’s RAM and may slow it down.  
The main DAQ Central window does not respond. A configuration window  
may be open. Configuration windows must be closed before continuing to  
interact with the software. Check your desktop and Windows taskbar for a  
Device Configuration window or an Options window.  
The device does not appear in the device list. Select “Detect Devices” from the  
Device menu to re-scan for connected devices. See also “Power” under “Basic  
Checklist” and “USB drivers…” in this section.  
Thermocouple inputs are noisy. If you experience noise from thermocouple  
wires, you can connect a 100K resistor from AxL to COM to eliminate some noise.  
7-1  
8
Service and Calibration  
Section 8 – Service & Calibration  
Your Omega DAQ components have been built and factory calibrated to meet or  
exceed the specifications listed here in this manual. The following section  
provides information on how to have your device serviced.  
8.1 Service & Calibration  
If any of your DAQ system components require service or calibration, please call  
our Customer Service Department at 1-800-622-2378 or 203-359-1660. They will  
assist you in arranging the return and service of your device. We can also be  
reached on the Internet at www.omega.com, e-mail: [email protected]  
8-1  
9
Specifications  
Section 9 – Specifications  
General Isolation:  
500V from PC  
External Excitation Output Voltage: +12v DC Regulated,  
Maximum total current output 67mA  
Power Requirements:  
Powered from USB port max 500 mA,  
or from external 7.5 to 12 Vdc,  
(1.0 A recommended)  
Thermocouple Types:  
J, K, T, E, R, S, B, N  
Cold-Junction Compensation  
Accuracy:  
1.0°C  
Isolation:  
500 Vdc from PC common  
Channel Capacities:  
Volts Inputs: 8 DE, or 16 SE  
Thermocouple Inputs: 8 DE  
* DE = Differential Mode,  
SE = Single ended mode  
Environmental:  
0 to 50°C (0 to 122°F)  
95% RH (non-condensing)  
Operating Temperature:  
0 to 50°C (32 to 122°F),  
0 to 95% RH non –condensing  
Analog Input Accuracy  
Differential Input:  
(exclusive of noise)  
0.015% of reading + .004% of range  
(In very slow mode)+ 10uV  
(exclusive of noise)  
Single-end Input:  
0.05% of reading + 0.01% of range  
(In very slow mode) + 50uV  
(exclusive of noise)  
Storage Temperature:  
Weight:  
-40 to 85°C (-40 to 185°F)  
0.5 lbs (0.23 kg)  
Dimensions:  
107 W x 128 L x 39mm H (4.2 x 5.1 x 1.5")  
Input Voltage Range:  
Software-programmable On A Per-  
channel Basis  
Differential/Single-ended:  
-10 to 10V  
-5 to 5V  
-2.5 to 2.5V  
-2 to 2V  
-1 to V -500 to 500 mV  
-250 to 250 mV  
-125 to 125 mV  
-75 to 75 mV  
-30 to 30 mV  
9-1  
9
Specifications  
Thermocouple Input Range:  
Type J: -18 to 1200°C (0 to 2192°F)  
Type K: -129 to 1372°C (-200 to 2502°F)  
Type T: -101 to 400°C (-150 to 752°F)  
Type E: -184 to 1000°C (-300 to 1832°F)  
Type R: 204 to 1768°C (400 to 3214°F)  
Type S: 204 to 1768°C (400 to 3214°F)  
Type B: 538 to 1820°C (1000 to 3308°F)  
Type N: -129 to 1300°C (-200 to 2372°F)  
Cold Junction  
Compensation Accuracy:  
1.0°C  
Input Voltage Range Relative  
To Analog Common (COM):  
Input Resistance:  
-10 to +10 Vdc  
1MΩ (single ended);  
2MΩ (differential); 5%  
Absolute Maximum Input Voltage:  
20 V maximum,  
relative to analog common  
Warm-up:  
1 hour to rated specifications  
USB 2.0 (full-speed)  
USB Device Type:  
Device Compatibility:  
Power Supply:  
USB 1.1, USB 2.0  
From USB(Port) or 9 Vdc adaptor  
External Power Source:  
Required when used with a bus-powered  
hub;  
DIN Rail Mounted For  
Rack Application:  
Optional  
Open Thermocouple Detect:  
Automatically enable when a channel is  
configured for a thermocouple sensor.  
Limitations of the Total Recording  
Memory Size:  
1 G or 500M (in PC)  
32 bit  
MCU:  
NOTE:  
All temperature specifications assume unit is held in relatively  
still air environment.  
NOTE:  
UNIV-AC-100/240, the external power supply is a Universal  
Power Adaptor. It is a 100-240 VAC to+9 Vdc output  
adaptor  
9-2  
9
Specifications  
Table 9-1. Thermocouple Accuracy (ºC) for OM-DAQ-USB-2401  
TC Type  
Temp.  
(°C)  
Typical  
Slow  
Fast  
-18  
0
1200  
1.3  
1.1  
0.9  
3.8  
3
2.5  
J
-129  
0
1.8  
1.4  
1.3  
5
3.9  
3.7  
K
T
E
R
1372  
-101  
0
400  
1.6  
1.4  
1
4.5  
4
2.9  
-184  
0
1000  
1.2  
0.9  
0.7  
3.4  
2.6  
1.9  
204  
700  
1768  
5.2  
4.6  
3.9  
14.8  
13.1  
11  
204  
700  
5.6  
5.2  
4.6  
16  
14.7  
12.8  
S
B
1768  
538  
1820  
7.9  
4.8  
13.7  
13.7  
-129  
0
1300  
2.6  
2
1.4  
7.3  
5.9  
4
N
NOTE:  
Thermocouple accuracy excludes cold junction compensation  
error.  
9-3  
10  
Approvals and Regulatory Compliance  
10.1 International Usage & CE Marking  
The OM-DAQ-USB-2401 multiple channel USB data acquisition module is CE  
marked and certified for use in several European countries. Please contact  
OMEGA for information on International Regulatory Compliance for each  
country.  
It is your (the user’s) responsibility to insure that these products are operated  
within the guidelines here in this manual and in conformance with all local or  
national regulations and laws.  
10-1  
NOTES:  
10-2  
NOTES:  
10-3  
NOTES:  
10-4  
WARRANTY/DISCLAIMER  
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a  
period of 13 months from date of purchase. OMEGA’s WARRANTY adds an additional one (1) month  
grace period to the normal one (1) year product warranty to cover handling and shipping time. This  
ensures that OMEGA’s customers receive maximum coverage on each product.  
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service  
Department will issue an Authorized Return (AR) number immediately upon phone or written request.  
Upon examination by OMEGA, if the unit is found to be defective, it will be repaired or replaced at no  
charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser,  
including but not limited to mishandling, improper interfacing, operation outside of design limits,  
improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of  
having been tampered with or shows evidence of having been damaged as a result of excessive corrosion;  
or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating  
conditions outside of OMEGA’s control. Components in which wear is not warranted, include but are not  
limited to contact points, fuses, and triacs.  
OMEGA is pleased to offer suggestions on the use of its various products. However,  
OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for any  
damages that result from the use of its products in accordance with information provided by  
OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by the  
company will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR  
REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF  
TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY  
AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF  
LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability of  
OMEGA with respect to this order, whether based on contract, warranty, negligence,  
indemnification, strict liability or otherwise, shall not exceed the purchase price of the  
component upon which liability is based. In no event shall OMEGA be liable for  
consequential, incidental or special damages.  
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic  
Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical  
applications or used on humans. Should any Product(s) be used in or with any nuclear installation or  
activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility  
as set forth in our basic WARRANTY/DISCLAIMER language, and, additionally, purchaser will indemnify  
OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the  
Product(s) in such a manner.  
RETURN REQUESTS/INQUIRIES  
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE  
RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN  
(AR) NUMBER FROM OMEGAS CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID  
PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return  
package and on any correspondence.  
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent  
breakage in transit.  
FOR WARRANTY RETURNS, please have the  
following information available BEFORE  
contacting OMEGA:  
FOR NON-WARRANTY REPAIRS, consult OMEGA  
for current repair charges. Have the following  
information available BEFORE contacting OMEGA:  
1. Purchase Order number under which the product  
was PURCHASED,  
1. Purchase Order number to cover the COST  
of the repair,  
2. Model and serial number of the product under  
warranty, and  
3. Repair instructions and/or specific problems  
relative to the product.  
2. Model and serial number of the product, and  
3. Repair instructions and/or specific problems  
relative to the product.  
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords  
our customers the latest in technology and engineering.  
OMEGA is a registered trademark of OMEGA ENGINEERING, INC.  
© Copyright 2012 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied,  
reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without the  
prior written consent of OMEGA ENGINEERING, INC.  
Where Do I Find Everything I Need for  
Process Measurement and Control?  
OMEGA…Of Course!  
Shop online at omega.comSM  
TEMPERATURE  
ߜ
 
ߜ
 
ߜ
 
ߜ
 
ߜ
 
Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies  
Wire: Thermocouple, RTD & Thermistor  
Calibrators & Ice Point References  
Recorders, Controllers & Process Monitors  
Infrared Pyrometers  
PRESSURE, STRAIN AND FORCE  
ߜ
 
ߜ
 
ߜ
 
ߜ
 
Transducers & Strain Gages  
Load Cells & Pressure Gages  
Displacement Transducers  
Instrumentation & Accessories  
FLOW/LEVEL  
ߜ
 
ߜ
 
ߜ
 
ߜ
 
Rotameters, Gas Mass Flowmeters & Flow Computers  
Air Velocity Indicators  
Turbine/Paddlewheel Systems  
Totalizers & Batch Controllers  
pH/CONDUCTIVITY  
ߜ
 
ߜ
 
ߜ
 
ߜ
 
pH Electrodes, Testers & Accessories  
Benchtop/Laboratory Meters  
Controllers, Calibrators, Simulators & Pumps  
Industrial pH & Conductivity Equipment  
DATA ACQUISITION  
ߜ
 
ߜ
 
ߜ
 
ߜ
 
ߜ
 
Data Acquisition & Engineering Software  
Communications-Based Acquisition Systems  
Plug-in Cards for Apple, IBM & Compatibles  
Data Logging Systems  
Recorders, Printers & Plotters  
HEATERS  
ߜ
 
ߜ
 
ߜ
 
ߜ
 
ߜ
 
Heating Cable  
Cartridge & Strip Heaters  
Immersion & Band Heaters  
Flexible Heaters  
Laboratory Heaters  
ENVIRONMENTAL  
MONITORING AND CONTROL  
ߜ
 
ߜ
 
ߜ
 
ߜ
 
ߜ
 
ߜ
 
Metering & Control Instrumentation  
Refractometers  
Pumps & Tubing  
Air, Soil & Water Monitors  
Industrial Water & Wastewater Treatment  
pH, Conductivity & Dissolved Oxygen Instruments  
M5025/0112  

Lexmark 13b0500 User Manual
Samsung Avmkc026ea B 0 User Manual
Sanyo Spw K095xh User Manual
Soleus Air Sg Wac 12ese F User Manual
Sony Found Not Kdl 52v4100 User Manual
Teledyne Automobile Parts 207 User Manual
Toshiba E Studio 55 User Manual
Xerox 5101 User Manual
York Ychbc User Manual
Zenoah G800bpu User Manual