Teledyne Welding Consumables 7500ZA User Manual

OPERATING INSTRUCTIONS FOR  
Model 7500ZA  
NDIR Infrared Gas Analyzer  
P/N M7500ZA  
4/09/14  
DANGER  
Toxic gases and or flammable liquids may be present in this monitoring system.  
Personal protective equipment may be required when servicing this instrument.  
Hazardous voltages exist on certain components internally which may persist  
for a time even after the power is turned off and disconnected.  
Only authorized personnel should conduct maintenance and/or servicing.  
Before conducting any maintenance or servicing, consult with authorized  
supervisor/manager.  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Specific Model Information  
It is not recommended that this instrument be used for analysis on  
any other gas or gas mixture than that specified at the time of purchase.  
The analyzer is set up and calibrated at the factory for a specific  
application using a known gas mixture that is representative of the  
customers’ process. Using this instrument to analyze any other gas  
mixture may result in interference and possible measurement error.  
Consult the factory for additional information for gas analysis not  
specified at the time of purchase.  
Instrument Serial Number: _______________________  
NO  
SO2  
CO2  
CO  
CH4  
O2  
Built-In  
External  
Measurable  
Species:  
Fuel Cell  
ZrO2  
Range 1:  
Range 2:  
Span Gas 1:  
Span Gas 2:  
Zero Gas:  
Background:  
Teledyne Analytical Instruments  
iii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 7500ZA  
Safety Messages  
Your safety and the safety of others is very important. We have  
provided many important safety messages in this manual. Please read  
these messages carefully.  
A safety message alerts you to potential hazards that could hurt you  
or others. Each safety message is associated with a safety alert symbol.  
These symbols are found in the manual and inside the instrument. The  
definition of these symbols is described below:  
GENERAL WARNING/CAUTION: Refer to the instructions  
for details on the specific danger. These cautions warn of  
specific procedures which if not followed could cause bodily  
Injury and/or damage the instrument.  
CAUTION: HOT SURFACE WARNING: This warning is  
specific to heated components within the instrument. Failure  
to heed the warning could result in serious burns to skin and  
underlying tissue.  
WARNING: ELECTRICAL SHOCK HAZARD: Dangerous  
voltages appear within this instrument. This warning is  
specific to an electrical hazard existing at or nearby the  
component or procedure under discussion. Failure to heed this  
warning could result in injury and/or death from  
electrocution.  
Technician Symbol: All operations marked with this symbol  
are to be performed by qualified maintenance personnel only.  
NOTE: Additional information and comments regarding a  
specific component or procedure are highlighted in the form  
of a note.  
No  
Symbol  
CAUTION:  
THE ANALYZER SHOULD ONLY BE USED FOR THE  
PURPOSE AND IN THE MANNER DESCRIBED IN  
THIS MANUAL.  
Teledyne Analytical Instruments  
iv  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
IF YOU USE THE ANALYZER IN A MANNER OTHER  
THAN THAT FOR WHICH IT WAS INTENDED,  
UNPREDICTABLE BEHAVIOR COULD RESULT  
POSSIBLY ACCOMPANIED WITH HAZARDOUS  
CONSEQUENCES.  
This manual provides information designed to guide you through  
the installation, calibration operation and maintenance of your new  
analyzer. Please read this manual and keep it available.  
Occasionally, some instruments are customized for a particular  
application or features and/or options added per customer requests.  
Please check the front of this manual for any additional information in  
the form of an Addendum which discusses specific information,  
procedures, cautions and warnings that may be peculiar to your  
instrument.  
Manuals do get lost. Additional manuals can be obtained from  
Teledyne at the address given in the Appendix. Some of our manuals are  
available in electronic form via the internet. Please visit our website at:  
www.teledyne-ai.com.  
Teledyne Analytical Instruments  
v
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 7500ZA  
Additional Caution on Safety  
To operate the analyzer properly, be sure to read “Caution on  
Safety” carefully.  
The descriptions listed here provide important information on safety.  
Be sure to observe them at all times. These safety precautions are classified  
into 3 levels: “DANGER,” “CAUTION” and “PROHIBITION”.  
Improper handling may cause dangerous  
situations that may result in death or serious  
injury.  
DANGER  
Improper handling may cause dangerous  
situations that may result in medium-level  
troubles, minor injury, or property damage.  
CAUTION  
Items which must not be done are indicated.  
PROHIBITION  
Caution on installation and transport of analyzer  
The unit is not of explosion-proof. Do not use it in an  
DANGER  
atmosphere of explosive gases. Otherwise, serious  
accidents including explosion or fire may result.  
For installation, observe the rule on it given in the  
CAUTION  
instruction manual, and select a place where the  
weight of analyzer can be supported. Installation in  
an inadequate place may cause turnover or falling,  
resulting in injury.  
Be sure to wear protective gloves when lifting the  
analyzer. Lifting it with bare hands may result in  
injury.  
Be sure to fix the cover before transporting the  
analyzer. Transportation in unstable state may result  
in injury.  
Teledyne Analytical Instruments  
vi  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
The gas analyzer is heavy. To transport the  
analyzer, please use a hand cart or equivalent. Avoid  
carrying analyzer by hand as much as possible.  
Doing so may cause injury.  
Take care not to let cable chips and other foreign  
objects enter the unit during installation work.  
Otherwise, fire, failure, or malfunction may result.  
Caution on piping  
Be sure to observe the following  
precautions while installing piping. Improper  
piping may result in gas leakage.  
DANGER  
If the leaking gas contains a toxic  
component, serious accidents may result. If  
it contains combustible gases, explosion or  
fire may result.  
Connect pipes correctly referring to the  
instruction manual.  
Discharge the exhaust gas outdoors to  
prevent it from remaining within the  
sampling device or indoors.  
Relieve the exhaust gas from the analyzer to  
the atmospheric pressure to prevent buildup  
of undesirable pressure to the analyzer.  
Otherwise, piping within the analyzer may be  
disconnected, resulting in gas leakage.  
Use pipes and pressure reducing valves to  
which no oil/grease is attached to the  
piping. Otherwise, fire may result.  
Caution on wiring  
Be sure to turn off the power before  
installing wiring. Otherwise, electric shock  
may result.  
DANGER  
Be sure to perform protective earth ground  
connection. Otherwise, electric shock or  
failure may result.  
Select a proper wiring material that satisfies  
Teledyne Analytical Instruments  
vii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 7500ZA  
the ratings of the instrument. Otherwise,  
electric shock or fire may result.  
Be sure to connect a power supply of  
correct rating. Otherwise, fire may result.  
Caution on use  
Be sure to read the instruction manual for  
reference gases before handling reference gases  
such as calibration gas to use them properly.  
DANGER  
CAUTION  
Leaving the analyzer unused for a long time or  
restarting it after long-term suspension requires  
procedures different from normal operation or  
suspension procedures. Be sure to follow the  
instructions in each instruction manual.  
Otherwise, intended performance may not be  
achieved. Also, accidents or injury may result.  
Do not operate the analyzer for a long time with  
its cover left open. Otherwise, dust, foreign  
matter, etc. may contaminate the internal walls,  
thereby causing faults.  
Do not touch the input/output terminals with  
metal or finger. Otherwise, electric shock or  
injury may result.  
PROHIBITION  
Do not smoke or use flames near the analyzer.  
Otherwise, fire may result.  
Do not allow water to enter the analyzer.  
Otherwise, electric shock or internal fire may result.  
Caution on maintenance and check  
Before performing work with the cover of the  
analyzer open for maintenance and check, be sure  
to purge completely not only within the analyzer but  
also measuring gas lines using nitrogen or air.  
Otherwise, poisoning, fire, or explosion may result  
due to gas leakage.  
DANGER  
Teledyne Analytical Instruments  
viii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Be sure to observe the following to perform work  
safely, avoiding electric shock or injury.  
CAUTION  
Remove the watch and other metallic objects before  
work.  
Do not touch the instrument with wet hands.  
If the fuse is blown, eliminate the cause and replace  
it with the one of the same capacity and type.  
Otherwise, electric shock or accidents may result.  
Do not use replacement parts other than those  
specified by the manufacturer. Otherwise, intended  
performance may not be achieved and accidents or  
failures may result.  
Dispose replacement parts such as maintenance  
parts as incombustibles according to the local waste  
disposal regulations.  
Teledyne Analytical Instruments  
ix  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 7500ZA  
Table of Contents  
Specific Model Information ......................................................... iii  
Safety Messages .......................................................................... iv  
Additional Caution on Safety ...................................................... vi  
List of Figures ............................................................................ xiv  
List of Tables............................................................................... xv  
Introduction ................................................................................... 1  
1.1 Overview  
1
2
1.2 Typical Applications  
Items included with the Analyzer................................................. 3  
2.1 Confirmation of delivered items  
2.2 Name and Description of Analyzer  
3
4
Installation ..................................................................................... 7  
3.1 Installation Conditions  
8
8
3.2 Installation  
3.2.1 Installation of Analyzer Main Frame  
3.3 Piping  
8
9
3.4 Sampling  
12  
12  
12  
12  
13  
13  
14  
15  
3.4.1 Sample Gas Conditioning  
3.4.2 Sample Gas Flow  
3.4.3 Preparation of Standard Gas  
3.4.4 Internal Purging of the Instrument Housing  
3.4.5 Pressure at Sample Outlet  
3.4.6 Example Configuration of Sample System  
3.5 Electrical Connections  
Operation ..................................................................................... 25  
4.1 Preparation for operation  
25  
25  
4.2 Warm-up Operation and Normal Operation  
Teledyne Analytical Instruments  
x
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Display And Operation Panels ...................................................27  
5.1 Front Panel Description  
27  
29  
30  
34  
5.2 Overview of Display and Operation Panels  
5.3 Outline of Display Screen  
5.4 Basic Operation  
Settings and Calibration .............................................................35  
6.1 Range Switch  
6.1.1 Setting the Range Switch Mode  
35  
35  
36  
36  
37  
37  
38  
40  
6.1.2 Manual Range Mode  
6.1.3 Range Identification Contacts  
6.2 Calibration Setting  
6.2.1 Setting the Span Gas Concentration  
6.2.2 Manual Zero Calibration  
6.2.3 Setting the Calibration Range  
6.2.4 Setting the Component/Range for Auto Calibration 42  
6.3 Alarm setting  
44  
44  
46  
47  
49  
49  
52  
52  
53  
55  
55  
57  
58  
58  
60  
68  
68  
6.3.1 Setting Alarm Parameters  
6.3.2 Hysteresis Setting  
6.3.3 Peak Alarm Setting  
6.4 Setting Up the Auto Calibration  
6.4.1 Auto Calibration  
6.4.2 Forced Run/Stop Of Auto Calibration  
6.4.2.1 Execution of auto calibration (only once)  
6.4.2.2 Forced stop of auto calibration  
6.5 Setting Auto Zero Calibration  
6.5.1 Auto Zero Calibration  
6.5.2 Forced Run/Stop of Auto Zero Calibration  
6.5.2.1 Execution of auto zero calibration (only once)  
6.5.2.2 Forced stop of auto zero calibration  
6.6 Parameter Setting  
6.7 Maintenance Mode  
6.7.1 Sensor Input Value screen  
Teledyne Analytical Instruments  
xi  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 7500ZA  
6.7.2 Error Log screen  
68  
6.7.3 Calibration Log screen  
6.7.4 Output Adjustment Screen  
6.7.5 Other Parameter  
69  
70  
71  
73  
73  
74  
6.8 Calibration  
6.8.1 Zero Calibration  
6.8.2 Span Calibration  
7. Maintenance ............................................................................ 75  
7.1 Daily Check  
75  
76  
76  
78  
78  
79  
80  
82  
82  
84  
7.2 Daily Check And Maintenance Procedures  
7.3 Long Term Maintenance  
7.4 Cleaning the Sample Cell  
7.4.1 Sample Cell Disassembly  
7.4.1.1 Pipe Cell Removal (See Fig. 7-1)  
7.4.1.2 Block Cell Removal (See Fig. 7-2)  
7.4.1.3 Removing the Measuring Unit  
7.4.2 Cleaning the Sample Cell  
7.5 Fuse Replacement  
8. Error Messages ....................................................................... 85  
8.1 Troubleshooting  
8.2 Error Log File  
86  
87  
9. Specifications.......................................................................... 89  
9.1 General Specifications  
9.1.1 Standard Specifications  
9.1.2 Standard Functions  
89  
89  
93  
9.1.3 Optional Functions  
94  
9.1.4 Performance  
98  
9.1.5 EC Directive Compliance  
9.1.6 Requirements for Sample Gas  
9.1.7 Installation Requirements  
98  
99  
100  
9.2 Table 1 Measurable Component and Range - Availability  
Check Table  
101  
Teledyne Analytical Instruments  
xii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
9.3 Code Symbols  
9.4 Outline Diagram  
106  
109  
Teledyne Analytical Instruments  
xiii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 7500ZA  
List of Figures  
Figure 2-1A: Model 7500ZA Front Panel......................................... 4  
Figure 2-1B: Model 7500ZA Rear Panel ......................................... 5  
Figure 7-1: Measuring Unit Configuration (Pipe Cell).................... 79  
Figure 7-2: Measuring Unit Configuration (Block Cell) .................. 81  
Figure 7-3: Measuring Unit with 2-Component Block and Pipe Cell83  
Teledyne Analytical Instruments  
xiv  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
List of Tables  
Table 7.1 Maintenance and Check Table......................................76  
Table 8-1: Error Messages............................................................85  
Teledyne Analytical Instruments  
xv  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 7500ZA  
DANGER  
COMBUSTIBLE GAS USAGE  
WARNING  
This is instrument is approved as an intrinsically safe gas  
analyzer for use in a category (ia) Group IIC hazardous area.  
This approval only to the equipment specified installed in  
accordance with the information contained within this  
manual. It is the customer's responsibility to ensure safety  
especially when combustible gases are being analyzed since  
the potential of gas leaks always exist.  
The customer should ensure that the principles of operating  
of this equipment is well understood by the user and that the  
instrument as well as any approved support equipment is  
properly installed. Misuse of this product in any manner,  
tampering with its components, or unauthorized substitution  
of any component may adversely affect the certification and  
the safety of this instrument.  
Since the use of this instrument is beyond the control of  
Teledyne, no responsibility by Teledyne, its affiliates, and  
agents for damage or injury from misuse or neglect of this  
equipment is implied or assumed.  
Teledyne Analytical Instruments  
xvi  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Introduction  
Introduction  
1.1 Overview  
The Model 7500ZA is designed to measure the concentration of  
NO, SO2, CO2, CO and CH4 in a sample gas based on the principle that  
different atomic molecules have a different absorption spectrum in the  
infrared band and the intensity of absorption is determined by the  
Lambert-Beer law.  
When equipped with the optional O2 sensor, the Model 7500ZA is  
capable of measuring up to 5 components simultaneously or up to 4  
components if O2 sensor is not installed.  
The Model 7500ZA is a microprocessor based analyzer with a large  
LCD display, superior accuracy and a multitude of functions suitable for  
applications such as:  
Combustible gas emissions from boilers or incinerators  
Steel gas analysis (blast furnace, steel converter, thermal  
treatment furnace, sintering (pellet equipment) etc.  
Coke furnace  
Storage and maturity of vegetable and fruit,  
Biochemistry (microbe) (fermentation)  
Air pollution monitoring [incinerator, exhaust gas  
desulfurization, denitration], automotive emission (excluding  
tester)  
Disaster prevention (detection of explosive gas and toxic gas,  
combustion gas analysis of new building material)  
Agricultural  
Chemical analysis (petroleum refinery plant, petroleum  
chemistry plant, gas generation plant)  
Environmental analysis  
Physical and chemical research  
Teledyne Analytical Instruments  
1
Download from Www.Somanuals.com. All Manuals Search And Download.  
Introduction  
Model 7500ZA  
1.2 Typical Applications  
The Model 7500ZA is a microprocessor based analyzer with a large  
LCD display, superior accuracy and a multitude of functions suitable for  
a variety of applications including:  
Combustible gas emissions from boilers or incinerators  
Steel gas analysis (blast furnace, steel converter, thermal  
treatment furnace, sintering (pellet equipment) etc.  
Coke furnace  
Storage and maturity of vegetable and fruit,  
Biochemistry (microbe) (fermentation)  
Air pollution monitoring (incinerator, exhaust gas  
desulfurization, denitration)  
automotive emission (excluding tester)  
Disaster prevention (detection of explosive gas and toxic gas,  
combustion gas analysis of new building material)  
Agricultural  
Chemical analysis (petroleum refinery plant, petroleum  
chemistry plant, gas generation plant)  
Environmental analysis  
Physical and chemical research  
Teledyne Analytical Instruments  
2
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Items Included with Analyzer  
Items included with the Analyzer  
2.1 Confirmation of delivered items  
The following items are included with you analyzer.  
Analyzer  
Fuse (2pcs))  
Std: IEC127-2  
05x20 mm  
250V/2A SB  
Analog Output Connector (1)  
25 pin D-sub male  
M2.6x4 mm  
Instruction Manual  
External Input Connector (1)  
(External O2/External Zirconia Sensor as  
specified)  
Digital Input Connector  
(3 max with the no. of DIO when digital  
I/O function is specified).  
Teledyne Analytical Instruments  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
Items Included with Analyzer  
Model 7500ZA  
RS-484 Connector  
(1) (when provided with communications  
function)  
9 pin D-sub connector male  
M2.6 x 4 mm  
Ferrite Core  
(1) for power cable (When terminal block  
for power supply is specified)  
Power Supply Cord (1) (when power input  
is specified)  
2.2 Name and Description of Analyzer  
The operator interface is shown in Figure 2-1A and the various  
connections to the instrument are shown in figure 2-1B.  
Figure 2-1A: Model 7500ZA Front Panel  
Teledyne Analytical Instruments  
4
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Operation  
Figure 2-1B: Model 7500ZA Rear Panel  
Teledyne Analytical Instruments  
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
Items Included with Analyzer  
Model 7500ZA  
[Blank page]  
Teledyne Analytical Instruments  
6
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Instalation  
Installation  
Installation of the Model 7500ZA NDIR Analyzer includes:  
Unpacking  
Mounting  
Gas connections  
Electrical connections  
Calibration  
DANGER:  
THIS UNIT IS NOT AN EXPLOSION-PROOF  
INSTRUMENT. DO NOT USE IT IN A PLACE WITH  
EXPLOSIVE GASES TO PREVENT EXPLOSION, FIRE  
OR OTHER SERIOUS ACCIDENTS.  
CAUTION:  
ENTRUST THE INSTALLATION, MOVEMENT OR RE-  
INSTALLATION TO A SPECIALIST OR THE  
SUPPLIER. A POOR INSTALLATION MAY CAUSE  
ACCIDENTAL TIPOVER, ELECTRIC SHOCK, FIRE,  
INJURY, ETC.  
The gas analyzer is heavy. It should be installed with utmost  
care. Otherwise, it may tip over or fall, causing accident or  
injury.  
For lifting the gas analyzer, be sure to wear protective gloves.  
Using bare hands may result in injury.  
This unit should be installed in an environment conforming to  
the conditions noted in the instruction manual. Otherwise,  
electric shock, fire or other malfunction of the unit may occur.  
During installation, care should be taken to keep the unit free  
from entry of chips, dust, and debris. Otherwise, it may cause  
fire, trouble or malfunction of the unit.  
The analyzer is shipped with all the materials you need to install  
and prepare the system for operation. Carefully unpack the transmitter  
and inspect it for damage. Immediately report any damage to the  
shipping agent.  
Teledyne Analytical Instruments  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
Installation  
Model 7500ZA  
3.1 Installation Conditions  
To install the analyzer for optimum performance, select a location  
that meets the following conditions;  
1. This instrument is designed for rack mounting in a standard 19”  
equipment rack.  
2. Indoor use only.  
3. Vibration-free.  
4. Free of dust and debris and obstruction around the analyzer.  
5. AC power:  
Operating voltage:  
Rated frequency:  
100V to 240 VAC  
85V to 264V AC  
50/60 Hz  
Power consumption: 100 VA max.  
6. Operating conditions:  
Ambient temperature : -5° to 45°C (max. 40°C when two optical  
units are used, and the power supply is more than 200V AC)  
7. Maintenance space:  
When analyzer is installed by itself, adhere to the  
diagram to make sure there is adequate space to  
remove the cover for maintenance.  
When the analyzer is installed as an unit, See  
Section 3.2.1  
8. A circuit breaker that meets IEC60947-1 and  
IEC60947-3 specifications should be installed.  
9. A breaker should be installed near the analyzer  
where an operator can access it.  
10. A label that clearly identifies the breaker should be visible.  
11. The breaker rating should meet the analyzer rating max 2A and  
conform to all necessary approvals.  
3.2 Installation  
3.2.1 Installation of Analyzer Main Frame  
Installation methods for the analyzer main unit is shown below.  
Teledyne Analytical Instruments  
8
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Instalation  
The analyzer weight must be supported at the bottom of the  
casing.  
The analyzer should be installed in a place where ambient  
temperature is within -5 to 45°C (max. 40°C when two optical  
units are used, and the power supply is more than 200V AC), and  
temperature fluctuation during operation is minimum.  
Where vibration is unavoidable, protect the analyzer from  
vibrating. If necessary, install rubber or other vibration absorbing  
material around the case to isolate any vibration from the support  
structure.  
3.3 Piping  
Observe the following items when making gas connections to the  
rear panel:  
Piping should be connected to the gas inlets and outlets at the  
rear panel of the analyzer.  
Use a corrosion resistant tubing comprised of Teflon, stainless  
steel or polyethylene to connect the instrument to a sampling  
system. Even if there is a danger of corrosion, refrain from using  
rubber or soft vinyl tubing. These materials will cause inaccurate  
indication due to gas absorption from the tubing.  
Pipe connection port fittings are NPT 1/4” female thread (or 1/4”  
RC). Piping should be cut as short as possible for optimum  
response. About 4 mm inner diameter is recommended.  
Teledyne Analytical Instruments  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
Installation  
Model 7500ZA  
Entry of dust into the instrument may result in poor performance.  
Use clean tubing and couplings.  
Sample inlet:  
Attach clean, dry sample gas to the inlet port.  
Note that the calibration gases ( zero and span)  
also connect to this port.  
Input gas flow t should be constant within the  
range of 0.5 L/min ±0.2 L/min.  
Sample outlet:  
Purge gas inlet:  
Attach tubing for exhausting sample or calibration  
gases. The exhaust line should be directed to an  
appropriate safe vent depending on the nature of  
the sample/calibration gas used.  
Used to connect a purge gas for purging the  
internal housing of the analyzer. Use dry N2 or  
instrument air for purge gas. (Flow rate is 1L/min  
or more. Keep dust or moisture/mist from  
entering the purge line.  
CAUTION:  
THIS INSTRUMENT IS DESIGNED TO HANDLE  
HAZARDOUS GASES. WHENEVER A FITTING IS  
OPENED, USE A NEW FERRULE AND CONE TO  
SECURE A GAS TIGHT SEAL. EACH FITTING MUST  
BE LEAK CHECKED WHENEVER A CONNECTION  
HAS BEEN OPENED OR DISTURBED IN ANY  
MANNER.  
The internal piping diagram for the Model 7500ZA is shown  
below.  
Teledyne Analytical Instruments  
10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Instalation  
Correspondence between measured components and the optical  
units is as follows:  
Measured  
Components  
Optical Bench 1  
Single component  
CO2/CO  
Optical Bench 2  
None  
Single measured species:  
NO, CO2, CO or CH4  
Two measured species:  
CO2, CO  
None  
Two measured species:  
NO/CO, NO/SO2  
NO/CO  
CO/SO2  
SO2/CO  
SO2/CO  
Three measured species:  
NO/ SO2/ CO  
NO  
Four measured species:  
NO/ SO2/ CO  
NO/CO  
Teledyne Analytical Instruments  
11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Installation  
Model 7500ZA  
3.4 Sampling  
3.4.1 Sample Gas Conditioning  
1. Dust from the sampling gas should be completely removed using  
a filter. For the final stage filter, use a filter that removes dust  
particles down to 0.3µm.  
2. Dew point must be lower than the ambient temperature. If vapor  
is contained in the sample, dew point should be lowered to 2°C  
using a dehumidifier.  
3. If SO3 mist is present in the sample gas, use a mist filter or  
cooler to remove the SO3 mist. Other mists should be removed  
using a mist filter or gas dryer.  
4. Corrosive gases such as Cl2, F2 and HCl, if present in  
considerable amounts, will shorten the life of component parts.  
5. Sample gas temperature should be within 0 to 50°C. Do not flow  
hot gas directly into the instrument.  
3.4.2 Sample Gas Flow  
Sample flow should be 0.5L/min ± 0.2L/min. Avoid any flow  
fluctuation during measurement. Observe the flow reading using the  
flowmeter provided as shown in the example of the sampling system  
configuration (Item 3.4.6).  
3.4.3 Preparation of Standard Gas  
Routine calibration is required once a week using span and zero  
calibration gases to keep the instrument performing optimally. Prepare a  
standard gas cylinder for zero calibration and span calibration per the  
table below.  
Without O2  
Sensor  
Built-in O2  
Sensor  
External O2  
Sensor  
Zero Gas  
N2  
N2  
Dry air  
(O2 of 99.9 to 100  
vol% for reverse  
range O2  
measurement.)  
Gas with  
concentration  
Gas with  
concentration of  
Gas with  
concentration of  
Span gas other  
than for O2  
Teledyne Analytical Instruments  
12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Instalation  
Without O2  
Sensor  
Built-in O2  
Sensor  
External O2  
Sensor  
measurement.  
of 90 to 100% on 90 to 100% on  
90 to 100% on  
the measuring  
range.  
the measuring  
range  
the measuring  
range.  
Gas with conc. of O2 gas at 1 to 2%  
90 to 100% on  
the measuring  
Span gas for O2  
measurement  
range or  
atmospheric air  
(21% O2).  
(O2 gas of 95 to  
95.5 vol% for  
reverse range O2  
measurement.)  
3.4.4 Internal Purging of the Instrument Housing  
The internal housing does need not be purged except under the  
following conditions:  
1. A combustible gas component is contained in the sample gas.  
2. Corrosive gas is contained in the atmospheric air at the  
installation site.  
3. The same gas as the sample gas component is contained in the  
atmospheric air at the installation site.  
In such cases, the analyzer housing should be purged with the  
instrument air or dry N2.  
The purge flow rate should be set to about 1L/min. The purge gas  
must be free of dust and moisture.  
3.4.5 Pressure at Sample Outlet  
Pressure at the sample gas outlet should be atmospheric pressure.  
Make sure there are no kinks or obstructions in the exhaust line.  
Teledyne Analytical Instruments  
13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Installation  
Model 7500ZA  
3.4.6 Example Configuration of Sample System  
The following illustrates a typical sample system configuration for  
five component gas measurement monitoring a combustion exhaust gas  
from boiler, refuse incinerator, etc.  
Contact TAI for other system configurations to match your specific  
application.  
Name  
Description  
Name  
Description  
Gas extractor with a  
heating type stainless  
steel filter of standard  
mesh 40µm.  
Adjusts and monitors  
the flow rate of the  
sample gas.  
(1) Gas extractor  
8) Flowmeter  
Removes drain, mist,  
and dust.  
The safety drain trap is  
divided into two spaces  
for positive and  
Reference gas used for  
calibrating zero and  
span of the analyzer,  
depending on the  
(2) Mist filter  
9) Reference gas  
(3) Safety drain trap  
measured gas.  
negative pressure. It  
monitors  
and adjusts the sample  
gas pressure.  
10) Zirconia O2  
sensor  
External zirconia  
oxygen sensor used for  
(4) Gas aspirator  
For aspiration of the  
sample gas  
Teledyne Analytical Instruments  
14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Instalation  
Name  
Description  
Name  
Description  
Dries the moisture in  
the sample gas to a  
dew point of approx.  
2°C  
Used for flowing the  
calibration gas.  
measuring the oxygen  
concentration in  
sample gas.  
(Not used when O2  
sensor is built-in.)  
(5) Electronic gas  
cooler  
(6) Solenoid valve  
(7) Membrane filter  
PTFE filter used to  
eliminate fine dust  
particles.  
Added to NOx Analyzer.  
A special catalyst  
material is used for  
efficient conversion of  
NO2 to NO  
(11) NO2/NO  
converter  
3.5 Electrical Connections  
CAUTION: To prevent electric shock or fire:  
Turn off the power before wiring.  
Always use a protective earth connection.  
Select a proper wiring material that satisfies the  
ratings of the instrument. • Be sure to connect a  
power supply of correct rating.  
The power terminal block and external input/output connector are  
installed on the rear panel. Refer to the diagram below.  
Teledyne Analytical Instruments  
15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Installation  
Model 7500ZA  
(1) Power Supply (standard terminal 1 to 2)  
Connect the power supply to the power terminal, and connect the ground  
wire to the grounding terminal (standard terminal 3). Be sure to provide  
a protective earth connection. Use solderless terminals (for M4) for  
connection to the terminals (power and earth). See the figure below.  
For the infrared gas analyzer, install an accessory ferrite core (to  
the power supply terminal stand side) on the AC power supply wiring  
line. See below.  
CAUTION:  
AFTER WIRING IS COMPLETE, BE SURE TO  
REPLACE THE PROTECTIVE COVER OVER THE  
TERMINAL BLOCKS.  
Note: Avoid installing this instrument near an electrical unit (high  
frequency furnace or electric welder) that generates  
electrical noise. If using the instrument near such a noise  
generating unit is unavoidable, use a different power line.  
Mount a noise suppressor such as varistor or spark  
quencher to the noise  
generating unit when  
source of noise is due to  
relays or solenoid valves.  
See accompanying figure.  
Mount the suppressor  
near the noise source;  
otherwise its effect will be  
diminished.  
Teledyne Analytical Instruments  
16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Instalation  
(2) Analog output signal: Analog output connector (A/O)  
Output signal :  
4 to 20 mA DC or 0 to 1 V DC (selected when  
ordering) Minus lines for the insulation and signal  
are common from the ground and internal circuit.  
Allowable load :  
4 to 20 mA DC, 550or less  
0 to 1 V DC, 100kor more  
The analog output signals of the instrument are not isolated  
individually. It is recommended to isolate the signals individually to  
eliminate the interference from unwanted signals or other external  
interference, especially if the cable exceeds 30 meters or leads to  
outdoors.  
(3) O2 Sensor Input: External input connector (A/I)  
Input signal:  
External zirconia O2 sensor: Zirconia O2 sensor signal (Fuji ZFK7  
output)  
External O2 sensor:  
0 to 1 VDC (DC input resistor of 1MW or  
more)  
Teledyne Analytical Instruments  
17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Installation  
Model 7500ZA  
Used when the external zirconia O2 sensor or the external O2  
sensor is ordered. It is not required or used if the analyzer has a  
built-in O2 sensor.  
Connect the dedicated connector (accessory) to the output of the  
external zirconia sensor or the external O2 sensor (shipped  
separately).  
When using an external O2 sensor, the input signal should be 0 to  
1 VDC with respect to the O2 full scale of the analyzer. The O2  
concentration display, output, and O2 correction can then be  
performed correctly.  
Note: The O2 sensor input is not isolated. TAI recommends that it  
be isolated when an external O2 analyzer is installed  
remote from the analyzer. The zirconia O2 sensor should  
be installed at a location that is as close to the analyzer as  
possible.  
The figure below shows how to connect the O2 signal to the  
dedicated connector (accessory).  
Teledyne Analytical Instruments  
18  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Instalation  
(4) Contact input/output (DIO): digital input/output connector  
(DIO 1 to 3)  
Contact input signal : Voltage is applied from the external 12 to 24 V  
DC, max 15mA.  
Photo-coupler isolation (from each DI and  
ground)  
Contact capacity :  
C contact relay output 24V/1A AC/DC  
resistive load  
Teledyne Analytical Instruments  
19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Installation  
Model 7500ZA  
Teledyne Analytical Instruments  
20  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Instalation  
Isolated output (from each DO and ground)  
To avoid external interference, the wiring from the analog output  
signal, O2 sensor input and contact input should be run separately from  
that of power supply and contact output.  
Note: Avoid excess noise generated by other nearby equipment,  
by grounding the analyzer and using properly shielded  
cables.  
(5) Communication: RS-485 connector  
(6) Timing of Contact Output for Calibration  
1. Manual calibration: (See “Item 6.8 Calibration” if the analyzer is  
equipped with the auto calibration feature).  
Teledyne Analytical Instruments  
21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Installation  
Model 7500ZA  
2. Automatic calibration:  
(See also Section 6.4: Automatic Calibration Settings)  
Teledyne Analytical Instruments  
22  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Instalation  
[Blank page]  
Teledyne Analytical Instruments  
23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Operation  
Operation  
4.1 Preparation for operation  
Tube and wiring check  
Double-check if tubes of the gas sampling and exhaust ports are  
correctly connected.  
Double-check for proper wiring.  
4.2 Warm-up Operation and Normal Operation  
(1) Operation Procedure  
1. Turn ON the power switch on the left side when facing the front  
panel of the analyzer unit. The measurement screen will appear  
on the front display panel in a few seconds.  
2. The instrument warm-up time is about 4 hours after which the  
instrument will produce an accurate measurement.  
Note: During the warm-up stage, the concentration reading may  
be above the upper range limit. When this occurs, the  
indication will display as a series of four dashes. However,  
this is not an error.  
3. During the warm up period, you can set the various instrument  
parameters as described in Section 6: Setting and Calibration.  
4. After the warm-up period, perform the zero/span calibration as  
described in Section 6.8 Calibration.  
5. After calibrating the analyzer, introduce sample gas into the  
analyzer and allow the lines to purge before accepting the  
measurement value.  
Teledyne Analytical Instruments  
25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Installation  
Model 7500ZA  
[Blank page]  
Teledyne Analytical Instruments  
26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Display and Operation Panels  
Display And Operation Panels  
This section describes the display unit and operation panel on the  
Model 7500ZA as well as the various functions available to the user.  
5.1 Front Panel Description  
Display unit:  
The measurement screen and the setting items are  
displayed.  
Operation panel: The operator interface panel is shown below.  
Teledyne Analytical Instruments  
27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Display and Operation Panels  
Model 7500ZA  
Name  
Description  
Name  
Description  
(1) MODE key Used to switch the mode.  
(5) ESC key  
Used to return to the  
previous screen or  
cancel the setting  
midway.  
Used to change the selected item (by  
(2) SIDE key  
(6) ENT key  
Used for confirmation of  
selected items or  
values, and for  
moving the cursor) and the numeral  
digit.  
execution of calibration.  
Used to change the selected item (by  
(3) UP key  
(7) ZERO key Used for zero  
calibration  
moving the cursor) and to increase the  
numeral value.  
Used to change the selected item (by  
moving the cursor) and to decrease  
the numeral value.  
(4) DOWN key  
(8) SPAN key Used for span  
calibration  
Teledyne Analytical Instruments  
28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Display and Operation Panels  
5.2 Overview of Display and Operation Panels  
Teledyne Analytical Instruments  
29  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Display and Operation Panels  
Model 7500ZA  
5.3 Outline of Display Screen  
(1) Measurement mode screen (appears when the power is  
turned on)  
The measurement screen depends on the number of components.  
The following screen configuration is shown as an example for NO,  
SO2, CO2, CO and O2 (output: 12 channels).  
Scroll the UP and DOWN keys to display more than 5 channels.  
No.  
Name  
Function  
(1)  
Component display  
Displays the component of instantaneous value, corrected  
instantaneous value, corrected average value, etc.  
(2)  
(3)  
(4)  
(5)  
Concentration display  
Range display  
Displays the measured value of concentration.  
Displays the range values.  
Unit display  
Displays the unit with ppm or mg/m3 and vol%.  
Displays the average time.  
Average time display  
Instantaneous value and concentration value:  
The concentration display of Ch (component) where sampling  
components such as “CO2”, “CO” and “O2” are displayed in the  
component display, are the current concentration values of the  
measured components contained in sample gas under test.  
Teledyne Analytical Instruments  
30  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Display and Operation Panels  
O2 corrected concentration values:  
Ch components in which “cv**” is displayed as “cv CO” in the  
display are calculated values derived from the following  
equation. Refer to Section 6.7 Maintenance Mode - Other  
Parameter.  
(21On )  
(21Os )  
C   
Cs  
where:  
On: The value of the O2 correction reference value (set by application)  
Os: Oxygen concentration (Vol%)  
Cs: Concentration of relevant measured component.  
Note that Os does not exceed the O2 limit value set in Section  
6.7 Maintenance mode - Other parameter.  
C: Sample gas concentration (O2 corrected).  
The corrected sampling components are NOX , SO2 and CO only.  
O2 corrected concentration average value:  
In the Ch (component) and O2 average value where “CV/AV **” is  
displayed as “ CV/AV CO” in the component display, a value obtained by  
averaging O2 corrected concentration value or O2 average value in a  
fixed time is output every 30 seconds.  
Averaging time can be changed between 1 to 59 minutes or 1 to 4  
hours according to the average time settings (See Section 6.7, Parameter  
Setting).  
(The averaging set time is displayed as “1h”, for instance, in the  
range display).  
* The measurement ranges for O2 correction concentration value  
and O2 correction concentration average value are the same as  
the measuring components. Also, the measurement range of the  
O2 average value is the same as for O2.  
Teledyne Analytical Instruments  
31  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Display and Operation Panels  
(2) Setting/Selection Screen  
Model 7500ZA  
The setting/selection screen is configured as shown below:  
In the status display area, the current display item is displayed.  
In the message display area, messages associated with operation  
are displayed.  
In the setting item and selection item display area, items or  
values to be set are displayed, as required. To navigate through  
the fields, move the cursor to any item using the UP, DOWN and  
SIDE keys.  
Teledyne Analytical Instruments  
32  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Display and Operation Panels  
(3) Contents of Measured Channel (Ch)  
The following table gives measurement channels and their contents  
according to the symbols.  
Teledyne Analytical Instruments  
33  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Display and Operation Panels  
Model 7500ZA  
5.4 Basic Operation  
Measurement mode  
The measurement mode can display up to  
5 channels in a single screen. If 5 channels or  
more are to be displayed in a single screen,  
press the UP or DOWN key to scroll the  
channels one by one.  
User mode displays  
Switch Ranges  
Calibration Parameters  
Alarm Setting  
Setting of Auto Calibration  
Setting of Auto Zero Calibration  
Parameter Setting.  
Press the UP or DOWN key to move  
the cursor to the desired function.  
Select the function by pressing the  
ENT (ENTER) key. The screen display  
will change to that of the selected  
function.  
Each function is described in Chapter  
6. Settings and Calibration.  
Teledyne Analytical Instruments  
34  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Settings and Calibration  
Settings and Calibration  
6.1 Range Switch  
6.1.1 Setting the Range Switch  
Mode  
The range switch has three modes:  
Manual (MR)  
Remote (RR)  
Auto Ranging (AR)  
To set the range switch mode:  
1. Press the MODE key while in  
measurement mode to display the  
User mode screen.  
2. Move the cursor to “Switch  
Ranges” and press the ENT key.  
3. In the “Channel Selection” screen  
that appears, move the cursor by  
pressing the UP or DOWN key,  
and select Ch (component).  
4. Then press the ENT key.  
5. The selected range switch mode  
will be highlighted. Press the UP  
or DOWN key to select the  
desired mode.  
6. Then press the ENT key to  
confirm the selection.  
7. If “MR” is selected, the cursor  
moves to “Range Switch.”  
Teledyne Analytical Instruments  
35  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Settings and Calibration  
Model 7500ZA  
6.1.2 Manual Range Mode  
The analysis range for the measured component can be switched  
manually as follows:  
1. (1) Select “MR” as range switch mode,  
and then press the ENT key.  
2. Using the UP/DOWN keys, move the  
cursor to the desired range selection.  
The highlighted arrow indicates the  
currently selectable range.  
3. Select it by pressing the ENT key, and  
analysis will be carried out in that  
range.  
Note: If “RR” or “AR” is selected as range  
switch mode, this operation cannot  
be performed.  
The ranges for O2 correction  
value,O2 correction average value,  
and O2 average value are  
automatically switched according to  
the instantaneous value range  
switch settings. (Same as for “RR”  
or “AR”.)  
To return to the previous screen, press the  
ESC key. If ESC is pressed in the middle of  
an entry or before the ENT key was pressed  
to accept a setting, it will reject the current  
selection and return you the display to the  
previous screen without making the changes .  
6.1.3 Range Identification Contacts  
The contacts for range identification on each channel are closed  
when Range 1 is active, and open when Range 2 is active.  
If the measurement value is held by a remote contact input or  
during calibration and range switch conditions are met, the contact status  
will change only after the hold condition is removed.  
Teledyne Analytical Instruments  
36  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Settings and Calibration  
6.2 Calibration Setting  
This mode is used to set  
calibration parameters such as span  
and zero gas concentration,  
calibration range, and auto  
calibration settings and range.  
When <User Mode> The  
“Calibration Parameters” screen that  
appears, the data shown at right is  
illustrated.  
6.2.1 Setting the Span Gas  
Concentration  
This screen allows you to set  
the cal gas concentrations for zero  
and span on each Ch used for  
calibration.  
1. Select < User mode > →  
< Calibration parameters >  
< Calibration value >.  
The “Calibration Value  
Settings” screen will appear.  
2. Select the Ch you want to  
change by pressing the UP or  
DOWN key. Then press the  
ENT key and cursor moves  
to the first editable parameter  
within that field.  
3. Select the concentration item  
you want to set (Zero or  
Span) by pressing the UP,  
DOWN or SIDE (>) key.  
The SIDE key only functions  
within the selected channel.  
Press the ENT key when the  
cursor is in the desired field.  
The first digit of the  
parameter is highlighted.  
Teledyne Analytical Instruments  
37  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Settings and Calibration  
Model 7500ZA  
4. Enter calibration gas concentration for zero and span. Use the UP  
or DOWN key to increment or decrement the displayed value  
one digit at a time. When the correct number is displayed for that  
digit, use the SIDE key to move to the next digit and repeat the  
increment/decrement process with the UP/DOWN keys.  
5. After setting, save the entry by pressing the ENT key. The saved  
value becomes valid with the next calibration event.  
Note: Enter calibration parameters settings for each range. If a  
zirconia O2 sensor is installed, select 21.00 for the Zero if  
air is used for calibration, and use the concentration listed  
on the cylinder for the span value.  
After saving the calibration values entered, close the Calibration  
Parameter screen by pressing ESC. This will return the display to the  
previous screen.  
Note: Pressing ESC during an entry (before pressing ENT) will  
abort the entry leaving the previous entry intact.  
Calibration parameter range of value  
NOx, SO2, CO2, CO, CH4, external and built in O2 sensor:  
Span:  
1-105% of full scale  
External O2 sensor:  
Zero:  
Span:  
0-25%  
0.1%-5% (vol.)  
Reverse Range:  
Zero:  
Span:  
100% (vol.)  
95.25-95.00% (vol.) O2  
The settings cannot be performed beyond the range limits.  
6.2.2 Manual Zero Calibration  
When setting the zero calibration in manual mode, you can set all  
components simultaneously or one by one.  
Teledyne Analytical Instruments  
38  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Settings and Calibration  
1. Select < User mode > < Calibration parameters > < Zero  
calibration >. The Zero  
Calibration screen will  
appears as shown.  
2. Select the Ch you want  
to change using the UP  
or DOWN key and then  
press the ENT key. The  
setting content becomes  
highlighted.  
3. Choose whether you  
want to set the zero  
values all together or  
individually by toggling  
“at once” or “each”  
with the UP or DOWN  
key. Each channel that  
is set to “at once” will be zero-calibrated at the same time.  
4. When “each” is selected, the specific Ch (component) can be  
selected and zero-calibrated individually. Press the ENT key  
after making the choice and the specified calibration is  
performed.  
To close the Manual Zero Calibration screen, press ESC. This will  
return the display to the previous screen.  
Note: Pressing ESC during an entry (before pressing ENT) will  
abort the entry leaving the previous entry intact.  
Example:  
Set each Ch (component) to either “each” or “at once” using the steps  
above.  
Setting “each”  
Select the Ch (component) on the manual zero calibration screen and  
then perform the zero calibration.  
Setting “at once”  
In a manual zero calibration, Ch (components) for which “at once”  
are selected will simultaneously be zero-calibrated.  
Teledyne Analytical Instruments  
39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Settings and Calibration  
Model 7500ZA  
6.2.3 Setting the Calibration Range  
This mode is used to set whether the channels use one or both of  
the calibration ranges.  
1. From the Calibration Parameters screen, select Calibration  
Range. The following screen appears.  
Teledyne Analytical Instruments  
40  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Settings and Calibration  
2. Select the Ch you want to change by pressing the UP or DOWN  
key, then press the ENT key. The setting content is highlighted.  
3. Toggle between “both” or “current” using the UP or DOWN key.  
If “both” is selected, zero or span calibration will be  
performed on both Range 1 and Range 2 on the selected Ch  
when the calibration is performed.  
If “current” is selected, zero or span calibration will be performed  
only for the range displayed when calibration is performed.  
4. Press the ENT key after the selection, and the specified  
calibration is performed.  
To close the Calibration Range Settings screen or to abort the entry  
midway, press the ESC key. The previous screen will return.  
Example:  
Ch1  
NOx  
Range 1: 0 to 200 ppm  
Range 2: 0 to 2000 ppm  
both  
Ch 2  
SO2  
Range 1: 0 to 200 ppm  
Range 2: 0 to 2000 ppm  
current  
Ch1: Range 1 and Range 2 are calibrated together  
Ch2: Only the currently displayed range is calibrated  
Note: When “both” is selected, the same calibration gas  
concentration must be used for both ranges.  
Teledyne Analytical Instruments  
41  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Settings and Calibration  
Model 7500ZA  
6.2.4 Setting the Component/Range for Auto Calibration  
In this section you can select the Ch (component) and the range to  
calibrate on for those components that have the auto ranging (AR)  
feature enabled.  
1. From the User Mode screen  
select < Calibration  
parameters > < Auto  
calibration component/range >.  
The “Auto Calibration  
Component Range” screen  
appears as shown.  
2. Select the Ch you want to  
change by pressing the or the  
key. Press the ENT key. The  
cursor will be highlighted in  
the selected channel.  
3. Select the range you want to  
use as the main calibration  
range by pressing the UP or  
DOWN key to move to the  
desired range.  
4. Press the ENT key, and  
subsequent calibrations will be  
performed on the selected  
range when either auto  
calibration or auto zero  
calibration is requested.  
Note regarding Auto Calibration  
Component/Range Selection  
The auto calibration or manual  
calibration of the component for  
which auto ranging (AR) is in effect  
will be performed on the range that is  
selected here. In this case, once the  
calibration is started, the range is  
automatically switched, and on  
completion of the calibration, the  
original range is resumed.  
Teledyne Analytical Instruments  
42  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Settings and Calibration  
The range identification contact is interlocked with the range after  
the switch. However, if the hold setting is set to “ON,” the contact status  
before calibration is maintained.  
5. Press the SIDE key in the state described in (3), and the  
highlight is toggled between “enable” and “disable” auto  
calibration.  
6. Select “enable” of “disable” by pressing the UP or DOWN key.  
7. Then press the ENT key to save your selection.  
To close the Auto Calibration Settings screen or to abort the entry  
midway, press the ESC key. The previous screen will return.  
Auto Calibration Operation  
Auto calibration is performed under the following rules.  
1. Zero calibration is performed at the same time, for the Ch  
(component) in which “enable” is selected at the time of auto  
calibration and auto zero calibration.  
2. Span calibration is performed in the order from smallest Ch No.,  
for the Ch (component) that have “enable” set in the Auto  
Calibrations Setting screen.  
Note: ZERO calibrations for components which have been set to  
“enable” in the Auto Calibration Settings screen (above)  
are performed in batch irrespective of the description in  
“6.2.2 Manual Zero Calibration.”  
Teledyne Analytical Instruments  
43  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Settings and Calibration  
Model 7500ZA  
6.3 Alarm setting  
6.3.1 Setting Alarm Parameters  
The High/Low limit concentration alarms  
are user adjustable. Five different alarm contact  
outputs can be used. The setting range is 0-  
100% full scale for each range.  
To change the alarm setting, set the alarm  
ON/OFF setting to OFF, and then change the  
value.  
1. Enter the “Alarm Setting” screen from  
the user mode. The screen shown on the  
right will display. Use the UP or DOWN  
keys to move the cursor to the desired  
Alarm No. or hysteresis. Then press  
ENT.  
2. After selecting the alarm the screen  
shown at the right will display. Use the  
UP or DOWN keys until the cursor is  
aligned with the desired item and press  
ENT.  
Note: Set the values so that H-limit value >  
L-limit value and that (H-limit value -  
L-limit value) > hysteresis.  
In the case where reverse range O2  
measurements are used set the  
values so that L-limit value > H-limit  
value.  
When “0” is set, the alarm is inactive.  
3. After setting the parameters, press ENT  
to save the settings.  
To close the Alarm Settings screen or to  
abort the entry midway, press the ESC key. The  
previous screen will return.  
Teledyne Analytical Instruments  
44  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Settings and Calibration  
Alarm Setting Descriptions  
The alarm contact number can be assigned to any channel.  
Channels can also have multiple alarms assigned to them.  
Channel:  
Channel setting targeted for alarm.  
H-Limit value:  
Sets the high limit value (concentration) for  
alarm.  
L-Limit value:  
Kind of Alarm:  
Sets the low limit value (concentration) for alarm.  
Select either High limit alarm, Low limit alarm,  
and High limit or Low limit alarm, HH limit  
alarm, and LL limit alarm.  
High, HH:  
Alarm contact closes when above  
H-limit alarm.  
Low, LL:  
Alarm contact closes when below  
L-limit alarm.  
High or Low: Alarm contact closes when above  
H-limit value or below lower limit  
value.  
ON/OFF:  
Enables the alarm function when set to ON, or  
disables it if set to OFF.  
Note: The H-limit value cannot be set below the L-limit value, and  
the L-limit value cannot be set above the H-limit value.  
If it is desired to set the H-limit value below the L-limit  
value already stored in the memory, reduce the L-limit  
value beforehand, and vice versa for setting the low limiot  
above the high limit.  
Note: After turning on power, the alarm logic trigger is inactive for  
10 minutes.  
When an H-limit alarm occurs, the  
“H-alarm” message displays in the relevant  
component field. Other alarm displays are:  
L-alarm for Low-limit alarm, HH-  
alarm for high limit alarm, HH-alarm for  
high-high alarm, and LL-alarm for low-  
low alarm.  
Teledyne Analytical Instruments  
45  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Settings and Calibration  
Model 7500ZA  
6.3.2 Hysteresis Setting  
The alarm hysteresis can be set or changed to prevent chattering of  
an alarm output about its alarm set point. To set or change the alarm  
hysteresis:  
1. Navigate to the “Alarm Setting” screen and  
then move the cursor to “Hysteresis” using  
the UP or DOWN key. Then press ENT to  
display the screen shown at the right.  
2. From this screen enter the hysteresis values  
by increasing or decreasing the value using  
the UP or DOWN keys. Move the cursor to  
the next digit with the SIDE key. When the  
value is correct, press ENT to save.  
The hysteresis can be set from 0-20% full  
scale on the selected range.  
To close the Hysteresis Settings screen or to  
abort the entry midway, press the ESC key. The  
previous screen will return.  
Note: The hysteresis is common to all alarms (components).  
Hysteresis in peak alarm setting described in Item 6.3.3  
should be set separately.  
Hysteresis (In case of upper limit alarm)  
An alarm output is turned ON if the measured value exceeds the  
upper limit value as shown below. Once the alarm output has been  
turned ON, it is not turned OFF until the indication falls below the  
hysteresis width from the upper limit value.  
Teledyne Analytical Instruments  
46  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Settings and Calibration  
6.3.3 Peak Alarm Setting  
When the peak number of times CO concentration exceeds the upper  
limit value during measurement reaches the set number, an alarm is provided.  
The peak alarm and this setting screen  
appear only when an option is added.  
1. Press the MODE key while in the  
Measurement mode, and the User  
mode appears.  
2. Using the UP or DOWN keys move  
the cursor to “Setting of Peak  
Alarm”. Then press ENT.  
3. In the “Peak Alarm Setting”  
selection screen that appears, point  
the cursor to any item you want to set  
using the UP or DOWN key. Then  
press ENT.  
4. Enter numeric values using the UP or  
DOWN keys. The setting will be  
saved after pressing the ENT key.  
Description of Setting Items  
Peak Alarm : ON/OFF of peak alarm  
Alarm Value : If measuring value exceeds  
the set alarm value, a peak  
counter increments by 1  
time.  
Alarm Count : When the alarm value is  
exceeded this many times per  
hour, the peak count alarm is  
activated (closed).  
Hysteresis :  
To prevent possible  
chattering when the  
measuring value may exceed  
the set peak concentration by  
only 1 time, the peak count  
has an allowance in the  
hysteresis width.  
Teledyne Analytical Instruments  
47  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Settings and Calibration  
Model 7500ZA  
Setting Range  
Alarm value: 10-1000 ppm  
Alarm count: 1- 99 times/hr  
Hysteresis : 0-20 % FS  
5 ppm step (initial value: 500 ppm)  
(initial value: 5 times)  
(initial value: 0% of full scale)  
[% FS] represents the percentage with the CO range regarded as  
100%.  
Action of Peak Alarm  
If the CO concentration exceeds the alarm value, counting will  
begin. If the number of peaks is over the set times per hour, a peak alarm  
contact output becomes closed (ON). If it is less than the set times per  
hour, it is open (OFF). Since 5 peaks /hour is exceeded in the cross-  
hatched area marked (1) in the above graph, the peak count alarm is  
turned ON. Since peaks of more than 5 per hour occur at the interval  
between (1) and (2), the peak count alarm remains ON whereas at (2),  
peaks are reduced to 4 per hour, and the alarm is turned OFF.  
Like the hysteresis of the alarm setting, the hysteresis prevents  
possible chattering when a measured gas fluctuates about the alarm  
value.  
* The peak alarm counting logic is not performed during the first 10  
minutes after the power is turned ON.  
Teledyne Analytical Instruments  
48  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Settings and Calibration  
Releasing peak count alarm  
To release the peak count alarm, set the peak alarm to OFF.  
Turning on the peak alarm initiates the count at 0.  
6.4 Setting Up the Auto Calibration  
6.4.1 Auto Calibration  
Auto calibration is automatically carried out at the time when zero  
span calibration are set. Before changing  
the setting of auto calibration, set the  
auto calibration function to OFF.  
1. Enter the “Setting of Auto  
Calibration” screen from the user  
mode, and the display shown at  
right appears. Use the UP and  
DOWN keys to align the cursor  
with the desired item and then  
press the ENT key.  
2. In the “Set Auto Cal” screen that  
appears, enter the values for the  
various settings. To make a value  
entry or change, use the UP and  
DOWN keys to increment or  
decrement the value on digit at a  
time. Use the SIDE key to move  
the cursor to the next digit.  
When all the settings are correct,  
press the ENT key to save them.  
Subsequent auto calibrations will  
use the saved settings.  
Description of Auto Cal Settings  
Start Time : Setting for the first  
calibration event (day of the  
week, hour, minute)  
Teledyne Analytical Instruments  
49  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Settings and Calibration  
Model 7500ZA  
Cycle : The period between the start time of one calibration and  
the next (unit: hour/day)  
Flow Time : The time required for purging with calibration gas.  
Time required for replacement of sample gas after the calibration  
is completed (Set by calibration gas. See the next page.)  
ON/OFF: Turns the auto calibration feature ON or OFF.  
To close the Auto Cal Settings screen or to abort the entry midway,  
press the ESC key. The previous screen will return.  
Gas flow time setting  
1. With the cursor on “Flow Time”, press the ENT key to bring up  
the Flow Time screen.  
2. Move the cursor to the gas  
you want to change by  
pressing the UP or DOWN  
key, and then press ENT.  
3. The highlighted value can be  
changed by pressing UP or  
DOWN key, and then using  
the SIDE key move the  
cursor to the right to alter the  
next digit.  
4. After changing the value,  
press the ENT key to save  
the setting.  
5. Press the ENT key again to  
return to the Auto Cal Setting screen.  
Note: Only the channels used are displayed on this screen. The  
Ex. time sets the output signal hold extension time that  
follows the completion of calibration. It is valid only when  
the hold setting is set to “ON.” The Ex. time set here is also  
the hold extension time used in a manual calibration.  
Teledyne Analytical Instruments  
50  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Settings and Calibration  
Auto calibration status contact output is closed during auto  
calibration (NO side), and is open in other cases.  
Example:  
Start Time  
Cycle  
Sun  
1
12:00  
day  
Flow Time  
Zero  
350 sec  
350 sec  
CH 1 Span  
CH 2 Span  
350 sec  
350 sec  
CH 3 Span  
CH 4 Span  
CH 5 Span  
300 sec  
300 sec  
300 sec  
Ex. Time  
ON  
ON/OFF  
Thus, with the above settings:  
Setting Range:  
Cycle:  
1-99 hours or 1-40 days (default value—7days)  
60-90 sec (default value—300 sec)  
Flow Time:  
Teledyne Analytical Instruments  
51  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Settings and Calibration  
Model 7500ZA  
Caution: When an auto calibration starts, the measurement  
screen appears automatically.  
Any operation other than “Stop Auto Calibration”  
(see Item 6.4.2) is not permitted during auto  
calibration. “Stop Auto Calibration” cannot be  
performed with the key lock to ON.  
To force a cancel of the auto calibration, set the  
key lock to OFF and then execute “Stop Auto  
Calibration”.  
Turn power OFF and then ON again at the time set  
as the next start time in auto calibration, and then  
repeat it in the set cycle.  
Remote Start  
Whether the auto calibration is set at ON or OFF, an auto  
calibration is always available via a remote start input.  
6.4.2 Forced Run/Stop Of Auto Calibration  
Auto calibration can be performed just once or forcibly stopped  
while the calibration is in process  
6.4.2.1 EXECUTION OF AUTO  
CALIBRATION (ONLY ONCE)  
1. In the “Set Auto Cal  
screen, move the cursor  
to “Auto Cal Run” using  
the UP or DOWN key,  
then press ENT.  
2. “Run” is highlighted,  
displaying a message to  
Teledyne Analytical Instruments  
52  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Settings and Calibration  
confirm the execution of auto calibration. Press the ENT key to  
execute the auto calibration or press the ESC key to cancel.  
6.4.2.2 FORCED STOP OF AUTO  
CALIBRATION  
This mode is used to stop the  
auto calibration forcibly.  
1. In the “Set Auto Cal” screens  
point the cursor to “Auto  
Calibration Stop” using the  
“UP or DOWN key. Then  
press ENT. (“Auto Calibration  
Stop” appears when this  
screen is selected while auto  
calibration is performed.)  
2. “Stop” is highlighted, displaying a  
message to confirm the stop of auto  
calibration. Press the ENT key to stop the  
auto calibration or press the ESC key to  
cancel the stop auto cal action.  
“Auto Calibration” screen  
Example  
When Ch 1 and Ch2 are enabled for auto  
calibration (see Section 6.2.4):  
Zero calibration  
A message, “Zero cal.” blinks at Ch1 and  
Ch2.  
Ch1 span calibration  
A message, “Span cal.” blinks at Ch1.  
Ch2 span calibration  
A message, “Span cal.” blinks at Ch2.  
Teledyne Analytical Instruments  
53  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Settings and Calibration  
Model 7500ZA  
Caution: During an auto calibration, all key operations are  
not permitted other than key lock ON/OFF and  
“Auto Calibration Stop.”  
When the key lock is ON, the “Auto Calibration  
Stop” is locked out as well.  
To stop “Auto Calibration” forcedly, set the key  
lock to OFF and then execute “Auto Calibration  
Stop.”  
Teledyne Analytical Instruments  
54  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Settings and Calibration  
6.5 Setting Auto Zero Calibration  
6.5.1 Auto Zero Calibration  
Zero calibration can be performed automatically using the settings  
applied in this screen.  
The channels for which auto calibration is to be performed are  
determined by the auto calibration settings described in Section 6.2.4.  
Before changing any of the auto zero calibration settings, set the  
auto zero calibration to OFF.  
1. Enter the “Set Auto Zero Cal”  
screen from the user mode. The  
screen shown at the right will  
appear.  
2. Use the UP or Down keys to  
move the cursor to the desired  
item and press the ENT key. In  
the “Set Auto Zero” screen that  
appears, set the desired  
parameter using use the UP or  
DOWN key to  
increment/decrement the value  
and the SIDE key to move to the  
next digit or field.  
After the setting is correct, press  
the ENT key to save the settings.  
The next auto zero calibration  
will use the entered values.  
Description of Auto Zero Settings  
Start Time : Setting for the first  
calibration event (day of the  
week, hour, minute)  
Cycle : The period between the  
start time of one calibration and  
the next (unit: hour/day)  
Teledyne Analytical Instruments  
55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Settings and Calibration  
Model 7500ZA  
Flow Time: The time required for purging with calibration gas.  
Time required for replacement of sample gas after the calibration  
is completed (Set by calibration gas. See the next page.)  
ON/OFF: Turns the auto calibration feature ON or OFF.  
To close the Auto Zero Settings screen or to abort the entry  
midway, press the ESC key. The previous screen will return.  
Auto Calibration Status Relay Contacts  
The auto calibration status contact output is closed during auto zero  
calibration (NO side), and open during normal operation.  
Example  
Start Time  
Cycle  
SUN  
12  
12:00  
hour  
sec  
Flow Time  
ON/OFF  
300  
ON  
For an auto zero calibration using the above settings:  
In this example, Ch1 through Ch5 are enabled, as shown in Section  
6.2.4 “Setting the Component/Range for AutoCalibration.  
Teledyne Analytical Instruments  
56  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Setting the Range  
Settings and Calibration  
Cycle  
1 to 99 hours or 1to 40 days (default: 7 days  
60 to 900 sec (default: 300 sec.)  
Flow Time  
Caution: When an auto zero calibration starts, the  
measurement screen automatically appears.  
Any operation other than “Auto Zero Calibration  
Stop” (see Item 6.5.2) is not permitted during auto  
zero calibration. “Auto Zero Calibration Stop”  
cannot be performed with the key lock set to ON.  
To forcibly cancel auto zero calibration, set the  
key lock to OFF and then execute “Auto Zero  
Calibration Stop”.  
If the auto calibration period and auto zero  
calibration period overlap, the auto calibration is  
retained, ignoring the auto zero calibration of that  
period.  
When the hold setting is set to ON, the hold time  
for auto calibration contact and measurement  
value output signal is extended after calibration  
for gas purging.  
Remote Start  
A remote start input will begin a calibration event regardless of  
whether the auto zero calibration is set to ON or OFF.  
6.5.2 Forced Run/Stop of Auto Zero Calibration  
Auto zero calibration can be performed just once, or auto zero  
calibration can be forcibly stopped during calibration.  
Teledyne Analytical Instruments  
57  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Settings and Calibration  
Model 7500ZA  
6.5.2.1 EXECUTION OF AUTO ZERO CALIBRATION (ONLY ONCE)  
1. In the “Setting of Auto Zero  
Calibration” screen that  
appears, point the cursor to  
“Run” by pressing the UP  
or DOWN key. Then press  
ENT.  
2. “Run” is highlighted,  
displaying a message to  
confirm execution of auto  
zero calibration. Press the  
ENT key to execute the  
calibration, or the ESC key  
to cancel.  
6.5.2.2 FORCED STOP OF AUTO ZERO CALIBRATION  
This mode is used to forcibly cancel an auto zero calibration event.  
1. In the “Setting of Auto Zero  
Calibration” screen that  
appears, point the cursor to  
“Stop” by pressing the UP  
or DOWN key, then press  
ENT. (“Auto Zero  
Calibration Stop” appears  
when the screen is selected  
while auto zero calibration  
is performed.)  
2. “Stop” is highlighted,  
displaying a message to  
confirm the stop of auto  
zero calibration.  
Press the ENT key to stop the auto zero calibration and the ESC  
key to cancel (not stopped).  
Teledyne Analytical Instruments  
58  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Settings and Calibration  
Auto Zero Calibration Screen  
Example  
With Ch1 and Ch2 set to enable for auto  
calibration (see Section 6.2.4):  
Zero calibration  
A message, “Zero cal.” blinks at Ch1  
and Ch2.  
Caution: During auto zero calibration,  
the only key operations allowed are key lock  
ON/OFF and “Auto Zero Calibration Stop.”  
When the key lock is set at ON, even the “Auto  
Zero Calibration Stop” cannot be used.  
To forcibly stop “auto zero calibration”, set the  
key lock to OFF and then execute “Auto Zero  
Calibration Stop.”  
Teledyne Analytical Instruments  
59  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Settings and Calibration  
Model 7500ZA  
6.6 Parameter Setting  
From the Parameter Setting screen you can set or adjust various  
parameters of the instrument such as time, key lock, etc. Settable items  
are as follows:  
Description of setting items  
Current Time : Current year, month, date, day of the week, hour,  
and minute setting. (The display appears in this order.)  
Note: The clock backup time is 2 days. If power is turned on  
after being off for 2 days or longer, the time must be set again.  
Key Lock : Invalidates any key operation except canceling the  
key lock.  
Output Hold : Sets whether Calibration Output is held or not, and  
the holding value setting.  
Response time : Sets the response time of electrical system.  
Average Period : Sets the moving average time.  
Backlight Timer : Sets automatic OFF of the backlight of display  
unit and its timer.  
Contrast : Adjusts contrast of the LCD.  
Maintenance Mode : A password protected entry screen used to  
access the instrument’s maintenance mode.  
* For the maintenance mode, see Section 6.7.  
1. Enter the “Parameter  
Setting” screen from the user  
mode, and the display shown  
at right appears. Use the UP  
or DOWN key to move the  
cursor to the desired item  
and then press ENT.  
Teledyne Analytical Instruments  
60  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Settings and Calibration  
2. In the “Parameter  
Setting” screen that  
appears, perform the  
value entry or the setting.  
For the value entry or  
setting change, use the  
UP or DOWN key, and  
the SIDE key to move  
the cursor to the right.  
To close the Parameter Setting  
screen or to abort the entry  
midway, press the ESC key. The  
previous screen will return.  
Setting Range  
Hold setting : 0 to 100% FS  
Response time : 1 - 60 sec. (default: 15 sec)  
Average period : 1 - 59 min or 1 - 4 hours (default value: 1 hour)  
1 - 59 minutes when the unit is set to minute and 1 - 4 hours  
when is set to hour.  
Backlight Timer : 1 - 60 min (default value: 5 min)  
Maintenance mode : 0000 to 9999 (default value: 0000)  
Output Hold  
Setting an output hold to ON causes the output signal of each  
channel to be held during the manual/auto calibration plus the purge  
flow time (see Section 6.4, Setting of Auto Calibration). However, an  
output signal can be held via an external input regardless of the Hold  
ON/OFF setting.  
Teledyne Analytical Instruments  
61  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Settings and Calibration  
Model 7500ZA  
Manual Calibration  
Auto Calibration  
Remote Input  
Screen Display During Holding  
A “Hold ON” message will blink on the measuring screen during a  
hold period.  
Teledyne Analytical Instruments  
62  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Settings and Calibration  
But during a manual calibration the screen displays the process of  
calibration, therefore a “Hold ON” message will not displayed even if  
the output signal is held, however the screen will display the hold  
extending time.  
If the calibration is cancelled after the calibration gas has been  
supplied, the holding extending time will be performed in either of  
manual or auto calibration operation.  
You can select the hold value from the value immediately before  
entering output hold. The choices are “current” or an arbitrary value  
input using “setting”.  
Use the procedures shown below for adjusting these settings.  
1. From the “Parameter  
setting” screen that  
appears, select “Output  
Hold”. Highlight either  
“ON” or “OFF” using the  
UP or DOWN key and  
then press ENT to select  
it and return to (1).  
2. When ON/OFF is  
highlighted, press the  
SIDE key. Toggle  
between “Current” or  
“Setting” using the  
UP/DOWN keys and then  
press ENT when the  
desired selection is  
highlighted.  
3. Pressing ENT when  
“Current” is selected will  
hold the value  
immediately before the  
hold was initiated and  
return the highlight to (1).  
Pressing the ENT key  
while “Setting” is  
selected will bring up the  
parameter hold screen where an arbitrary hold value can be set.  
Teledyne Analytical Instruments  
63  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Settings and Calibration  
Model 7500ZA  
4. On the parameter hold screen that  
appears, move the cursor to the Ch  
(component) you want to hold using  
the UP or DOWN key and then press  
ENT.  
5. A highlighted value indicates that it  
can be changed. Adjust the  
highlighted value with the UP or  
DOWN key and then move the  
cursor to the right to change the next  
digit using the SIDE key.  
6. After the value is changed, press the  
ENT key.  
Note: The setting is expressed as 1/1  
full scale range for both  
respective ranges.  
When 0 to 1000 ppm is selected  
as the range, and 10% FS is  
selected as hold setting, the  
output equivalent to 100 ppm is  
held irrespective of the measured  
value at that time.  
7. Press the ESC key to return to the  
parameter setting screen.  
Teledyne Analytical Instruments  
64  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Settings and Calibration  
Setting Description  
The instantaneous measurement value that is displayed cannot be  
held. (Output only can be held.)  
Optional modbus communications “Measurement concentration”  
register values are held.  
If set value is selected for hold, instantaneous O2 correction  
value is calculated and held based on the set value.  
Range identification contact output cannot be switched even if  
the range is switched during the hold.  
Response time  
The response time of the electrical system can be changed. Setting  
is available by components.  
Note: It does not provide exact  
seconds for the setting  
time, rather, it provides a  
guide of the setting time.  
The setting value can be  
modified as requested  
by the customer.  
Average Period  
This screen allows you to set an averaging period for the average  
values of O2 correction and O2 average.  
It enables you to set an  
average time of 1 to 59 minutes  
(1-minute step) or 1 to 4 hours  
(1-hour step).  
Changing the setting also  
resets the averaging of O2  
correction and O2 average value.  
(Pressing the ENT key resets  
averaging only for the  
components whose setting were  
changed.)  
Teledyne Analytical Instruments  
65  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Settings and Calibration  
Model 7500ZA  
Average Value Reset  
This mode is used to clear all average values, i.e. O2 correction  
average and O2 average, and restarts averaging. All average values are  
reset simultaneously. The indication value and output value is 0 ppm,  
vol% or so at the time of the reset input (based on average period  
settings).  
Example:  
If the average period was set to 1 hour:  
Sampling occurs every 30 seconds.  
Every 30 seconds, the average for the last 1 hour (time setting) is  
output.  
At the instant of resetting, zero is assumed for all past values.  
This means that the average value will not be correct for 1 hour  
after resetting.  
Backlight Timer  
Automatic OFF setting of the backlight of the LCD unit is user  
adjustable. When the specified time elapses during measurement screen  
display, the backlight automatically switches off. Pressing any key will  
turn the backlight back ON.  
Teledyne Analytical Instruments  
66  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Settings and Calibration  
On the Parameters screen, when ON is selected for the Backlight  
Timer function, the time until auto OFF will also display. Press the  
SIDE key when ON is  
highlighted and the time setting  
can be changed. Use the UP or  
DOWN key to change the time  
value and then press ENT to  
save the value.  
When OFF is selected, the  
backlight always remains on  
and does not automatically turn  
off.  
Contrast  
The LCD Contrast is user  
adjustable. With the contrast  
function highlighted, use the UP  
or DOWN keys to move the  
slider representing the contrast.  
When satisfied, press the ENT  
key to save the setting.  
Maintenance Mode  
Enter the password and then press the ENT key to enter the  
maintenance mode. The password can be set using the password setting  
screen from Other Parameter/Password Set function while in the  
maintenance mode. The next section describes in detail the functions  
available in maintenance mode.  
The default password is set at the factory to “0000.” You can enter  
the maintenance mode with this value before changing the password in  
maintenance mode to one of your choosing.  
Teledyne Analytical Instruments  
67  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Settings and Calibration  
Model 7500ZA  
6.7 Maintenance Mode  
This mode is used to check sensor input values,  
display of error log or setting of passwords, etc. This is  
a password protected area. To access the maintenance  
mode, enter the correct password from the Parameter  
Setting/To Maintenance screen.  
1. Select the Maintenance Mode from the  
Parameter Setting screen to display the  
Password Setting screen.  
2. Enter the password, and the Maintenance Mode  
item selection screen will be displayed. Point  
the cursor to the item you want to set by  
pressing the UP or DOWN key and then press  
ENT.  
Note: “To Factory Mode” is for TAI’s service  
engineers only.  
3. Press the ESC key to return to the Maintenance  
Mode item selection screen from each screen.  
6.7.1 Sensor Input Value screen  
There are 5 possible inputs on the Model 7500ZA:  
1. Input 1 to 4 : NDIR sensor digital value  
2. Input 5 : O2 sensor digital value  
6.7.2 Error Log screen  
This screen displays the error history and logs the  
last 14 instrument errors.  
See Section 8.1 Error Message for specific  
information regarding error number and its meaning.  
To clear an error from the log, select Clear Error  
Log and press the ENT key. The specific entry will be  
removed from the list.  
Teledyne Analytical Instruments  
68  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Settings and Calibration  
6.7.3 Calibration Log screen  
Calibration history can be displayed from this screen. Information  
available includes sensor input value, concentration value, and the date  
when zero/span calibration was performed for the last 10 calibration  
events. They are logged according to component.  
To clear an event, move the  
cursor to Clear Calibration Log and  
press the ENT key, and the  
calibration log is cleared completely.  
Other items appearing on the  
Calibration Log screen are:  
Z1 : Zero calibration (Z) of Range 1  
S1 : Span calibration (S) of Range 1  
Cnt : Value of measuring detector at  
the time of calibration  
Con : Concentration value displayed  
before calibration  
YDMH: Time of calibration event  
Teledyne Analytical Instruments  
69  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Settings and Calibration  
Model 7500ZA  
6.7.4 Output Adjustment Screen  
The Analog Output Adjustment screen allows the user to adjust the  
0-1 V and 4-20 mA output signal.  
Connect a digital multi meter to the specific output terminal  
requiring adjustment and adjust the value so that 4 mA or 0 V is output  
at zero concentration and 20mA or 1V is output at  
full scale or span.  
1. Move the cursor using the UP or DOWN  
key to the output (OUT No. and zero/span)  
to be adjusted, and then press the ENT key.  
The selected value is highlighted.  
2. Adjust the value, while watching the  
output, using the UP or DOWN key. When  
the correct value appears, press the SIDE  
key to select the next digit.  
3. On completion of the adjustment, press  
ENT to save the adjustment.  
Caution: Be sure to perform zero / span  
calibration when the range  
setting is changed. Otherwise,  
the measurement value may not  
be output properly.  
Teledyne Analytical Instruments  
70  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Settings and Calibration  
6.7.5 Other Parameter  
The Other Parameter screen hosts 5 user settable functions:  
Password Setting  
O2 Reference Value  
O2 Reference Value Limit  
Station Number  
Instrument Range Setting  
To change any of these items:  
Press the UP or DOWN  
key to move the cursor to  
the item whose setting is to  
be changed.  
The values for password,  
oxygen correction, limit, and station No. are highlighted.  
Press the UP or DOWN key to change the value. Use the SIDE  
key to move to the next digit to the right. Press the ENT to save  
the new value.  
Password Set  
The Password Set screen is used to change the password that is required  
move from the parameter setting screen to the maintenance mode. An  
arbitrary 4-digit number can be set to replace the default password for  
this purpose.  
Note: Record the password and store in a safe place. Otherwise  
you cannot enter the maintenance mode.  
O2 ref. Value and Limit  
This function sets the oxygen concentration reference value used in  
the oxygen correction calculation*. It is settable in the range from 00 to  
19%. The O2 Limit can also be set from this screen. This sets the  
oxygen concentration limit at the time of oxygen correction calculation.  
It is settable in the range from 01 to 20%.  
* Refer to the O2 correction concentration value in “5.3 Outline of  
display screen” for oxygen correction calculation procedure.  
Teledyne Analytical Instruments  
71  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Settings and Calibration  
Model 7500ZA  
Station No.  
This function allows the user to set the station No. for MODBUS  
communication. It is settable in the range from 00 to 32.  
Range Setting  
This function brings up a screen where the user can set the  
range for which concentration measurements are made. There  
are two settable ranges for each component each with a  
minimum and maximum value as requested at the time of  
purchase. The range to be used can be selected as Range 1 or  
Range 2.  
To set the range:  
1. Move the cursor to the item to be set using the UP or  
DOWN key, and then press ENT.  
2. Move the cursor to the Ch (component) whose setting is  
to be changed using the UP or DOWN key, and then  
press ENT.  
3. Move the cursor to the item whose setting is to be  
changed using the UP or DOWN key, and then press  
ENT.  
The value for range 1 and range 2 must fall within the  
MIN and the MAX range of the instrument (including  
the MIN and the MAX range), and range 1 must be less  
than range 2.  
4. Press the UP or DOWN key to change the value. Use  
the SIDE key to move to the next digit to the right.  
Press the ENT to save the new value. The measurement  
unit cannot be changed.  
5. The decimal point can also be changed when it is  
highlighted. Use the UP or DOWN key to change the  
placement of the decimal point.  
6. After all changes have been entered, press the ENT key  
to save.  
Teledyne Analytical Instruments  
72  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Settings and Calibration  
6.8 Calibration  
6.8.1 Zero Calibration  
This function is used for the zero  
point adjustment. A proper zero gas,  
suitable for the application, should be  
used. Refer to section 3.4 “Sampling”.  
1. From the Measurement screen,  
press the ZERO key to display the  
Manual Zero Calibration screen.  
2. Select the Ch (component) to be  
calibrated using the UP or DOWN  
key and then press the ENT key,  
and zero gas will be admitted to  
the analyzer.  
Note: For the Ch (components)  
where “both” is set in the  
calibration range setting (see  
section 6.2.3), zero calibration  
is carried out on both ranges.  
3. Wait until the indication stabilizes  
on the zero gas then press ENT.  
The zero calibration will be carried  
out on the highlighted range(s).  
Note: On Ch (components) for which  
“AR” is set (see Section 6.1.1  
Setting Range Switch Mode),  
the cursor automatically moves  
to the next range selected in  
“Setting of Auto Calibration  
Component/Range” (6.2.4),  
and calibration is carried out  
on that range.  
To close the Zero Calibration screen  
or to abort the entry midway, press the  
ESC key. The previous screen will return.  
Teledyne Analytical Instruments  
73  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Settings and Calibration  
Model 7500ZA  
6.8.2 Span Calibration  
The Model 7500ZA requires periodic span calibration at or near the  
full scale value on the range of interest. It requires a prepared standard  
gas containing a known concentration of the measured  
species. For span calibration gas on the NOX, SO2, CO2,  
CO and CH4 channels, TAI recommends using a gas source  
containing 90-100% of the measured species in a  
background known to be compatible with the instrument.  
For the oxygen channel, with units that include a built-in  
O2 sensor, use a prepared standard gas with an O2 concentration  
of 90 to 100% full scale on the anticipated measuring range.  
If the instrument is fitted with an external zirconia  
sensor, use a prepared standard of 1-2% O2.  
To perform a span calibration:  
1. Press the SPAN key on the Measurement screen to  
display the Manual Span Calibration screen.  
2. Select the channel (component) to be calibrated  
using the UP or DOWN key and then press ENT.  
Calibration gas will then flow to the sensor.  
Note: For the Ch (components) in which “both” is set  
in the calibration range setting (Refer to Section  
6.2.3), span calibration is performed on both  
ranges.  
3. Wait until the indication is stable. After the  
indication has stabilized, press ENT. Span  
calibration on the selected range will be performed.  
Note: On channels for which “AR” (auto ranging) is  
enabled (See Section “6.1.1) the cursor  
automatically moves to the range that is  
enabled. See Section 6.2.4 “Setting of auto  
calibration component/range”. Calibration is  
then carried out on that range.  
To close the Span Calibration screen or to abort the  
entry midway, press the ESC key. The previous screen will  
return  
Teledyne Analytical Instruments  
74  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Maintenance  
7. Maintenance  
7.1 Daily Check  
1. Zero calibration and span calibration  
1. Perform zero calibration. For the calibration procedures, refer to  
Section 6.8.1 Zero calibration.  
2. Following the zero calibration, perform a span calibration. For  
the calibration procedures, refer to Section 6.8.2 Span  
calibration.  
3. Zero/span calibration should be carried out once a week, or as  
required.  
2. Flow rate check  
1. Sample and purge gas flow rates are as follows:  
Sample gas flow :  
Purge gas flow :  
0.5L/min ± 0.2L/min  
About 1L/min  
2. Instrument checks and maintenance should be carried out every  
day, as required.  
Teledyne Analytical Instruments  
75  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
Model 7500ZA  
7.2 Daily Check And Maintenance Procedures  
Table 7.1 Maintenance and Check Table  
7.3 Long Term Maintenance  
Create a long-term maintenance component procurement plan  
based on the “Gas analyzer annual inspection plan” indicated below.  
Gas Analyzer Annual Inspection Plan  
The recommended replacement period of components varies  
depending on the installation conditions.  
1. The recommended replacement period is a recommended  
standard criterion and varies depending on the environment of  
the field, conditions of measuring gas and other factors.  
2. The recommended replacement period is not the warranty period.  
It is provided as a preventative maintenance program baseline  
schedule.  
Teledyne Analytical Instruments  
76  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Maintenance  
Installation Conditions  
1. Ambient temperature: -5°C to +40°C  
2. Humidity: 90%RH or less  
3. Corrosive gases: None  
4. No radiated heat, direct sunlight or rain/wind  
5. Dust: No more than local environmental standards permit  
6. Vibration: None  
Sample gas conditions  
1. Temperature: +60°C to +800°C  
2. Pressure: -3 to +3 kPa  
3. Moisture content: 30% or less  
4. Dust: 0.1 g/Nm3 or less  
5. Components: 0 to 500 ppm NOX, 0 to 500 ppm SO2, 0 to 2000  
ppm CO, 5% to 15% CO2, 0% to 21%O2, 0 to 100 ppm HCl,  
residue N2  
Please consult TAI Customer Service regarding gas analyzer  
maintenance service requirements.  
Annual Inspection Plan Sheet  
Teledyne Analytical Instruments  
77  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
Model 7500ZA  
7.4 Cleaning the Sample Cell  
Entry of dust or water in the sampling cell contaminates the interior  
of the cell resulting in drift. If dirty or contaminated, it must be cleaned.  
Check the sampling device, especially the filter, to prevent the cell from  
being dust or mist contamination.  
Caution: Maintenance should be performed only by  
properly trained and qualified personnel.  
In addition to the following steps, the user must  
adhere to any local and regional regulations.  
7.4.1 Sample Cell Disassembly  
There are two kinds of sampling cells used in the Model 7500ZA:,  
block cells (cell length: 4 mm, 8 mm, 16 mm, 32 mm) and pipe cells  
(cell length: 64 mm, 125 mm, 200 mm and 250 mm).  
A two-component analyzer may incorporate both sampling cells in  
one optical unit. In this case, detach the pipe cell and then block cell (see  
Fig. 7-3).  
Teledyne Analytical Instruments  
78  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Maintenance  
7.4.1.1 PIPE CELL REMOVAL (SEE FIG. 7-1)  
1. Stop all gas flow. If necessary, purge the pipe cell thoroughly  
with zero gas.  
2. Turn OFF the power switch and disconnect the AC power cord.  
3. Remove the cover (six screws).  
4. Remove the tube connected to the pipe cell.  
5. Loosen and remove the screw (No. 7) holding the cell retainer  
(No. 11) at both ends of the pipe cell.  
6. Remove the cell from the assembly and unscrew the infrared  
transmission window (No. 14) at both ends.  
7. For assembly, reverse the disassembly procedure.  
Figure 7-1: Measuring Unit Configuration (Pipe Cell)  
Teledyne Analytical Instruments  
79  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
Model 7500ZA  
7.4.1.2 BLOCK CELL REMOVAL (SEE FIG. 7-2)  
For steps 1 to 4, see 7.4.1.1, Pipe Cell Removal.  
5. Disconnect and remove detector output cables from detector  
output circuit board (No.12). Applying identification mark on  
top of removed cable connector will ensure proper pin  
assignment later.  
6. Unscrew the two screws (No. 10) that hold the detector to the  
light source unit to remove the detector from the measuring unit.  
The block cell can be removed together with the detector.  
7. To remove the block cell, unscrew the two screws (No. 6)  
holding the block cell to the detector. The infrared transmission  
window (No. 8) is just sandwiched (not fixed) between the  
detector and block cell. Keep the detector facing up, when  
removing this window.  
8. For assembly, reverse the disassembly procedures.  
Note: The O-ring (No. 9) is placed between the window holder  
and block cell. Use care in positioning the O-ring. With a  
two-component analyzer, install the two-component  
detector last. Make sure that no space is left between the  
one-component and two-component detectors. When  
inserting the detector output cable connector into the PCB  
board, be careful to attach the connector with proper pin  
assignment (top/bottom).  
Teledyne Analytical Instruments  
80  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Maintenance  
Figure 7-2: Measuring Unit Configuration (Block Cell)  
Teledyne Analytical Instruments  
81  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
Model 7500ZA  
7.4.1.3 REMOVING THE MEASURING UNIT  
Refer to Fig. 7-3. For step 1 to 4, see 7.4.1.1, Pipe Cell Removal.  
5. Disconnect and remove detector output cables from detector  
output circuit board (No.12). Applying identification mark on  
top of removed cable connector will ensure proper pin  
assignment later.  
6. Disconnect wiring to the 2-pin terminals of the IR light source  
assembly and chopper motor pin connector from the PCB (No. 17).  
7. Detach the six screws (No. 16) fastening the base plate (No. 3) to  
remove the measuring unit. .  
8. For assembly, reverse the disassembly procedures.  
Note: Special care should be taken when assembling or  
disassembling the measuring cell to avoid using excessive  
force to the detector pipe or light source unit pipe. If the pipe  
is deformed or damaged by excessive force, there is a  
potential for gas leaks and this, in addition to creating a  
potential toxic hazard, will result in inaccurate measurement.  
7.4.2 Cleaning the Sample Cell  
1. To clean the sample cell or IR transmission window, first clear  
away any dirt using a soft brush and then lightly wipe using a  
soft cloth. Do not use abrasive or paper towels.  
Note: The window is fragile. Handle with care. Avoid scratching  
the window. Do not rub the dirt from the window roughly.  
2. If the window or the sample cell interior is very dirty, use a soft  
line-free cloth moistened with absolute alcohol.  
3. If the window is corroded, rub off any scale from the window  
lightly with a soft cloth to which chrome oxide powder is  
applied. If it is excessively corroded, it should be replaced.  
4. When the cleaning operation is complete, assemble according to  
the sample cell assembly/disassembly procedures. Assemble the  
pipe carefully. If it becomes bent or damaged, replace it with a  
new part.  
5. Do not wash the sample cell components with water.  
Teledyne Analytical Instruments  
82  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Maintenance  
Figure 7-3: Measuring Unit with 2-Component Block and Pipe Cell  
Teledyne Analytical Instruments  
83  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
Model 7500ZA  
7.5 Fuse Replacement  
1. Turn “OFF” the main power supply switch to the analyzer.  
2. Rotate the fuse holder cap (shown in the figure above)  
counterclockwise and pull it out. The cap (fuse holder) will  
contain the fuse. Remove the blown fuse from the holder and  
replace it with a new one (250VAC/2A, Slo-Blo).  
3. Reinstall the fuse holder cap, turn ON the power supply switch.  
The analyzer will power up normally.  
Teledyne Analytical Instruments  
84  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Error Messages  
8. Error Messages  
Table 8-1 lists the error messages that may occur onscreen along  
with the probable cause for the error.  
Table 8-1: Error Messages  
When errors No. 1 to No. 3 and No. 10 occur, analyzing block error  
contact output closes. When errors No. 4 to No. 9 occur, the calibration  
error contact output closes.  
Teledyne Analytical Instruments  
85  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Error Messages  
Model 7500ZA  
8.1 Troubleshooting  
When error No. 1 occurs, remove the top cover of the analyzer and  
check the LED on the light source PCB. If the LED is off, then the error  
is caused by a light source disconnection.  
Errors 1-3 and10 indicate an analyzer failure or fault. Contact TAI  
for assistance.  
Errors 4-8 indicate that the calibration procedure may be incorrect.  
Check the following items, and if the error still occurs, contact TAI.  
1. Is the calibration gas flowing to the analyzer?  
2. Does the calibration operation match the supplied gas? (For  
example, zero calibration is performed while flowing the span  
gas.)  
3. Does the supplied gas concentration match the gas concentration  
set in the calibration concentration setting?  
Also, if errors 5 and 7 occur, you can perform calibration forcibly,  
following the procedure. This is used as a fault recovery when  
calibration fails and calibration contents are invalid or wrong.  
Typical Error Screen Display  
For Errors 1-4, 6, and 8-10:  
Teledyne Analytical Instruments  
86  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Error Messages  
Press the ECS key to delete the error display.  
If the ESC key is pressed without removing the cause of an error,  
the error will be displayed again.  
When more than one error occurs, use the SIDE key to move to  
the next error display.  
For errors 5 and 7:  
8.2 Error Log File  
When an error occurs, the history is saved in an error log file. The  
error log file is accessible in the maintenance mode.  
Teledyne Analytical Instruments  
87  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Error Messages  
Model 7500ZA  
Error Log Screen  
* Up to 14 errors can be saved in the error history; the oldest error will  
be deleted one by one every  
time a new error occurs.  
* If the power supply is turned OFF, the contents in the error log file  
will not be lost or damaged.  
Deleting Error History  
Press the ENT key on the above screen, and the “Clear Error Log”  
will be highlighted. Pressing ENT again will clear the error history.  
Teledyne Analytical Instruments  
88  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Specifications  
9. Specifications  
9.1 General Specifications  
9.1.1 Standard Specifications  
Measurement Method:  
NO, SO2, CO2, CO, CH4:  
Non-dispersion infrared-ray absorption  
method  
Single light source and single beams (single  
beam system)  
O2:  
Fuel cell O2 analyzer (built-in) or  
paramagnetic O2 analyzer (built-in) or  
zirconia O2 analyzer (externally installed  
TYPE: ZFK7)  
Measured Gas/Range:  
Min Range  
0-200 ppm  
0-200 ppm  
0-100 ppm  
0-200 ppm  
0-500 ppm  
0-10 vol.%  
Max Range  
0-5000ppm  
0-10 vol.%  
0-100 vol.%  
0-100 vol.%  
0-100 vol.%  
0-25 vol.%  
NO  
SO2  
CO2  
CO  
CH4  
O2  
(built in)  
Fuel Cell  
O2  
(built in)  
Paramag.  
0-5 vol.%  
None  
0-100 vol.%  
100-95 vol.%  
Teledyne Analytical Instruments  
89  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Specifications  
Model 7500ZA  
O2  
0-5 vol.%  
0-25 vol.%  
External  
Zirconia  
Max. 5 components measurement  
including O2.  
For reverse range O2 measurement, IR gas  
measurement is not available (single range  
O2 only).  
Measuring range ratio max. 1:10 (except  
O2)  
User adjustable ranges between the  
specified minimum and maximum range.  
Settable one range or two ranges, low  
range (Range 1) and high range (Range 2).  
See Table 1 for possible combinations of  
components and ranges.  
Display Indication:  
Digital indication in 4 digits (LCD panel  
with LED back light)  
Instantaneous value of each component  
Instantaneous value after O2 correction  
(only in NO, SO2, CO measurement with  
O2)  
Average value after O2 correction (only in  
NO, SO2, CO measurement with O2)  
O2 average value  
Analog Output Signals:  
4 to 20mA DC or 0 to 1V DC, isolated  
internally from circuit and ground. Output  
lines are non-isolated; 12 outputs max.  
Allowable load 550for 4 to 20mA DC  
Teledyne Analytical Instruments  
90  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Specifications  
Allowable load 100Kfor 0 to 1V DC  
Refer to Table2 for the channel No. of  
displayed values and analog output  
signals.  
Analog Input Signal:  
For signal input from externally installed  
O2 sensor.  
Signal requirement:  
(1) Signal from Zirconia O2 sensor  
(TYPE: ZFK7)  
(2) 0 to 1V DC from an O2 sensor. Input  
section is not isolated. This feature is  
present when an O2 sensor is not built  
in.  
*
Externally installed O2 sensor must be  
purchased separately.  
Digital Output: (option)  
1C contact (24V DC/1A, resistive load)  
max.15 outputs  
Instrument error, calibration error,  
range identification, auto calibration  
status, solenoid valve drive for auto  
calibration, High/Low limit alarm  
contact output.  
All relay contacts are isolated mutually  
and from the internal circuit.  
Digital Input: (option)  
Voltage contact (supply 12-24VDC  
(15mA Max.)) Max. 9 inputs  
Remote range change over, auto  
calibration remote start, remote hold,  
average value reset, Isolated from the  
internal circuit with photocoupler.  
Teledyne Analytical Instruments  
91  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Specifications  
Model 7500ZA  
Power Supply:  
Voltage rating:  
Allowable range:  
Frequency:  
100V to 240V AC  
85V to 264V AC  
50Hz/60Hz  
Power consumption: 100VA max.  
Operating Conditions:  
Ambient temperature: -5˚C to 45˚C  
(40˚C max. when 2  
optical system at  
200V AC power  
source)  
Ambient humidity:  
90% RH max.,  
non-condensing  
Storage Conditions:  
Ambient temperature: -20˚C to 60˚C  
Ambient humidity: 100% RH max.,  
non-condensing  
Dimensions (H × W × D): 133 x 483 x 382mm  
Mass:  
Approx. 9 kg  
Enclosure:  
Steel casing, for indoor use  
Material for Wetted Parts:  
Gas inlet/outlet; SUS304  
Sample cell; SUS304,chloroprene rubber  
Infrared-ray transmitting window: CaF2  
Internal piping: Toaron, Teflon,  
Polypropylene  
Paramagnetic O2 sensor cell : SUS316  
Fuel cell O2 sensor cell: ABS resin  
Gas inlet/outlet:  
Rc1/4 or NPT 1/4 internal thread  
1L/min ( when required)  
Purge gas flow rate:  
Life time of Fuel Cell O2 Sensor: 2 years  
Teledyne Analytical Instruments  
92  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Specifications  
9.1.2 Standard Functions  
Output Signal Hold:  
Output signals are held unchanged during  
manual and auto calibrations when HOLD  
function is ON.  
The values held are those just before start  
of calibration mode or setting value.  
User selectable.  
Indication of instantaneous values will not  
be held.  
Range Switching:  
Range switch function is available in  
manual, auto, and remote modes. Only  
preset switch method is effective.  
Manual: Allows range to switch by key operation.  
Auto: Automatically switched from first range to  
second range when the measured value  
exceeds 90%FS of first range.  
Automatically switched from second  
range to first range when the measured  
value drops to 80% or less first range.  
Remote (option): Voltage contact input allows the range to  
switch via an external signal when a  
remote range switch input is received.  
When the contact input terminals for each  
component are input voltage, the first  
range is selected, and it is switched to the  
second range when the terminals are open.  
* The switch range values are settable  
between the first range and second range  
values (low/high range values).  
Teledyne Analytical Instruments  
93  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Specifications  
Model 7500ZA  
9.1.3 Optional Functions  
Remote Output Holding:  
Output signal is held at the last value or  
preset value by a voltage input to the  
remote output holding input terminals.  
Holding is maintained while the voltage is  
input to the terminals. Indication of  
instantaneous values are not held.  
Range Identification Signal:  
The present measuring range is identified  
by a contact position.The contact output  
terminals close for each component when  
the first range is selected, and open when  
the second range is selected.  
Auto Calibration:  
Auto calibration is carried out periodically  
at the preset cycle. When a standard gas  
cylinder for calibration and a solenoid  
valve for opening/closing the gas flow line  
are installed externally by the customer,  
calibration will be carried out using the  
solenoid valve drive contacts for zero and  
span calibration and will be cycled on/off  
sequentially at the set auto calibration  
timing.  
Auto Cal Cycle Setting: Auto calibration cycle is user defined.  
Setting is variable within 1 to 99 hours (in  
increments of 1 hour) or 1 to 40 days (in  
increments of 1 day).  
Gas Flow Time Setting: The time for flowing each calibration gas  
in auto calibration is user defined and  
settable between 60 to 900 seconds (in  
increments of 1 second).  
Auto Cal Remote Start:  
Auto calibration starts by opening the auto  
cal remote start input terminal after short  
Teledyne Analytical Instruments  
94  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Specifications  
circuiting for 1.5 sec or longer. Auto  
calibration initiates when contacts open.  
Auto Zero Calibration:  
Auto zero calibration is carried out  
periodically at the preset cycle. This cycle  
is independent from “Auto Calibration”  
cycle.  
When zero calibration gas and solenoid  
valve for opening/closing the calibration  
gas flow line are installed by the customer,  
zero calibration will be carried out at the  
set auto zero calibration timing.  
Auto Zero Cal Cycle Setting: Auto zero calibration cycle is user settable.  
Setting is variable within 1 to 99 hours (in  
increments of 1 hour) or 1 to 40 days (in  
increments of 1 day).  
Gas Flow Time Setting: The timing for flowing zero gas in an auto  
zero calibration is user settable from 60 to  
900 seconds (in increments of 1 second)  
High/Low Limit Alarm:  
Alarm contact output turns on when  
measurement value reaches the preset high  
or low limit alarm value. Contacts close  
when the instantaneous value of each  
channel exceeds the high alarm limit value  
or falls below the low alarm limit value.  
Instrument Error Contact Output: Contacts turn on when analyzer  
error No. 1, 2, 3 or 10 occurs.  
Calibration Error Contact Output: Contacts turn on at occurrence of  
manual or auto calibration errors from 4-9.  
Auto Cal Status Contact Outputs: Contacts turn on during auto  
calibration.  
O2 Correction:  
Correction of measured NO, SO2 and CO  
gas concentrations into values at reference  
O2 concentration  
Teledyne Analytical Instruments  
95  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Specifications  
Model 7500ZA  
Correction formula:  
21On  
21Os  
C   
Cs  
C : Sample gas concentration after O2  
correction.  
Cs : Measured concentration of sample gas.  
Os : Measured O2 concentration (Limit setting:  
1 to 20% O2)  
On : Reference O2 concentration (value  
changeable by setting 0 - 19% O2).  
Average value after O2 correction and O2 average value  
calculation:  
The result of O2 correction or instantaneous  
O2 value can be output as an average value  
over the preset period of time.  
Moving average method is used. Sampling  
interval is 30 seconds (Output is updated  
every 30 seconds.  
Update is the averaged value of the most  
recent elapsed averaging time period.)  
Averaging time period is settable from 1-59  
minutes (in increments of 1 minute) or 1-4  
hours (in increments of 1 hour).  
Average Value Resetting:  
The above-mentioned output of average  
value is started from the initial state by  
opening the average value resetting input  
terminals after short circuiting for 1.5 sec  
or longer.  
Output is reset by input voltage and the  
terminal circuit.  
Communication Function:  
RS-485 (9pins D-sub connector)  
Half-duplex bit serial  
Teledyne Analytical Instruments  
96  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Specifications  
Start-stop synchronization  
ModbusTM protocol  
Contents : Read/Write parameters  
Read measurement conc. and  
instrument status  
Remark : When connecting via RS-232C  
interface, an RS-232C  
RS485 converter should be  
used.  
Atmospheric Pressure Correction:  
Measure atmospheric pressure and  
calculate compensation (for use, exhaust  
port of analyzer must be at atmospheric  
pressure.  
After atmospheric pressure correction:  
Zero point: Not affected.  
Span point: Change is 0.5% measured  
value or less relative to a 1%  
change in atmospheric  
pressure.  
Correction range: 700hPa-1050hPa  
Teledyne Analytical Instruments  
97  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Specifications  
Model 7500ZA  
9.1.4 Performance  
Repeatability:  
Linearity:  
±0.5% of full scale  
1% of full scale prior to atmospheric  
pressure correction (option)  
Zero drift:  
±2% of full scale/week.  
For auto zero calibration use 500ppm or  
lower range.  
Span drift:  
±2% of full scale/week  
Response time (for 90% FS response):  
1 to 15 sec electrical response. Within 10-  
30 seconds including purge time of sample  
gas.  
Gas replacement time depends on the  
number of measuring components and  
measuring range.  
Interference from other gases:  
9.1.5 EC Directive Compliance  
This product conforms to the requirements of the Low Voltage Directive  
2006/95/EC and EMC directive 89/336/EEC (as amended by Directive  
92/31/EEC), both as amended by Directive 93/68/EEC.  
It conforms to following standards for product safety and  
Teledyne Analytical Instruments  
98  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Specifications  
Electromagnetic Compatibility ;  
EN61010-1 : 2010 Safety requirements for electrical  
equipment for measurement, control and  
laboratory use.  
“Installation Category II”  
“Pollution Degree 2”  
EN61326-1 : 2006  
Electrical equipment for measurement,  
control and laboratory use — EMC  
requirements.  
9.1.6 Requirements for Sample Gas  
Flow rate:  
Temperature:  
Pressure:  
0.5L / min ±0.2L / min  
0 to 50˚C  
10 kPa or less (Gas outlet side should be at  
atmospheric pressure).  
Dust:  
100 µg/Nm3 or less in particle size of 0.3  
µm or less.  
Mist:  
Not allowed  
Moisture:  
For sample gases NO, SO2, CO(0-200  
ppm range): less than 2°C saturation point.  
For most other sample gases: less than  
standard room temperature saturation  
point.  
Corrosive component:  
1 ppm or less  
Standard gas for calibration:  
1) Infrared-ray measurable component, standard O2  
Zero gas: Dry N2  
Span gas ; Each sample gas having  
concentration 90 to 100% of its  
measuring range recommended).  
For external zirconia O2 sensor use the  
same calibration gas line:  
Teledyne Analytical Instruments  
99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Specifications  
Model 7500ZA  
Zero gas ; Dry air or atmospheric air (Do  
not use with CO2 measurement)  
Span gas ; For other than O2 measurement,  
each sample gas having  
concentration 90 to 100% of its  
measuring range.  
For O2 measurement, O2 gas of  
1 to 2 vol% /remainder N2 gas  
2) Reverse range O2 measurement  
Zero gas; 100vol% O2  
Span gas ; For O2 measurement, O2 gas of  
95.0%~95.5vol%/ remainder  
N2 gas  
*For reverse range O2 measurement,  
infrared measurable component is not  
measurable.  
9.1.7 Installation Requirements  
Indoor use (Select a place where the equipment does not receive  
direct sunlight, draft/rain or radiation from hot substances. If  
such a place cannot be found, a roof or cover should be prepared  
for protection).  
Avoid sites where unit receives heavy vibration.  
Select a site where atmospheric air is clean  
Teledyne Analytical Instruments  
100  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Specifications  
9.2 Table 1 Measurable Component and Range -  
Availability Check Table  
Procedure for range selection  
One component analyzer:  
1. First determine 1st range, then select 2nd range from the  
corresponding right column.  
More than two components analyzer:  
1. The 2nd range in the tables for two and more components is  
maximum available range.  
2. Select the 2nd range less than or equal to the “2nd range (max)”.  
Teledyne Analytical Instruments  
101  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Specifications  
Model 7500ZA  
Teledyne Analytical Instruments  
102  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Specifications  
Teledyne Analytical Instruments  
103  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Specifications  
Model 7500ZA  
Teledyne Analytical Instruments  
104  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Specifications  
Teledyne Analytical Instruments  
105  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Specifications  
Model 7500ZA  
9.3 Code Symbols  
Teledyne Analytical Instruments  
106  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Specifications  
note1) When “D” is specified at 4th digit, Power supply cord is supplied  
in the scope of supply. Cord specification should be specified at  
the 20th digit.  
note2) When only O2 measurement is necessary, “Y” should be  
specified at 6th digit.  
note3) When “1” is specified at 7th digit, O2 pt analyzer signal has to be  
set as 0-1V DC linear corresponding to full scale.  
External zirconia O2 analyzer and external O2 analyzer are not  
included in the scope of supply, and has to be separately ordered.  
note4) Refer to Table 1 for possible combination of measuring  
components and ranges in this manual. When “Y” is specified at  
6th digit, “Y” should be specified at 9th to 16th digit. For fuel  
cell O2 analyzer, range is 0-10vol% or more.  
note5) Select the type of voltage rating, plug type and applicable  
standard of the power supply cord by 20th digit. Select a power  
supply cord for using at the location of end-user. When “A” is  
specified at 4th digit, Power supply cord will not be included in  
scope of delivery. When English is specified for display  
language, “E” should be selected at 20th digit.  
note6) O2 correction is calculated only for NO, SO2 and CO.  
note7) When 5 components measurement is specified, “H” must not be  
specified at 22nd digit.  
Teledyne Analytical Instruments  
107  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Specifications  
Model 7500ZA  
When 4 components measurement is specified and “H” is  
specified at 22nd digit, 3 points is maximum for alarm output  
function.  
note8) When “B” is specified at 24th digit, measuring range should be  
specified by ppm range code. In this case NO, SO2 and CO  
measuring range are corresponding range in mg/m3. Please refer  
to the table shown below for the corresponding range code based  
on “mg/m3”.  
note9) When A to D is specified at 25th digit, the analyzer will be  
adjusted and delivered with the following gases.  
Standard “A”: balance gas N2.  
For heat treatment furnace “C”: balance gas 30vol%  
H2/remaining N2.  
For converter “D”: balance gas CO, CO2.  
When other adjustment is required, please specify “Z”.  
When “Z” is specified, please attach a list of gas composition  
contained in the measuring gas.  
Teledyne Analytical Instruments  
108  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NDIR Gas Analyzer  
Specifications  
9.4 Outline Diagram  
<Analyzer main unit>  
Teledyne Analytical Instruments  
109  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Specifications  
Model 7500ZA  
[Blank page]  
Teledyne Analytical Instruments  
110  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Telex Stereo Amplifier 2225 User Manual
Thomas Betts Power Supply Limitorque MX Device Net Field Unit User Manual
Toastmaster Convection Oven RM 5T User Manual
Toastmaster Mixer 1770U User Manual
Toro Smoke Alarm PS90 User Manual
Toshiba Computer Drive HDD1544 User Manual
Toshiba Modem PCX2000 User Manual
Tripp Lite Power Supply SUIOK User Manual
Tripp Lite TV Cables EZA 050 User Manual
Ultimate Support Systems Microphone PRO SB User Manual