Teledyne Photo Scanner M5000B User Manual

OPERATING INSTRUCTIONS  
Model 5000B  
Photometric Analyzer  
DANGER  
HIGHLYTOXICANDORFLAMMABLELIQUIDSORGASESMAYBEPRESENTINTHISMONITORINGSYSTEM.  
PERSONALPROTECTIVEEQUIPMENTMAYBEREQUIREDWHENSERVICINGTHISSYSTEM.  
M5000B  
HAZARDOUSVOLTAGESEXISTONCERTAINCOMPONENTSINTERNALLYWHICHMAYPERSISTFORA  
TIMEEVENAFTERTHEPOWERISTURNEDOFFANDDISCONNECTED.  
ONLYAUTHORIZEDPERSONNELSHOULDCONDUCTMAINTENANCEAND/ORSERVICING. BEFORE  
CONDUCTINGANYMAINTENANCEORSERVICINGCONSULTWITHAUTHORIZEDSUPERVISOR/MANAGER.  
11/02/2006  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Photometric Analyzer  
Table of Contents  
Part I: Control Unit ................................Part I: 1-1  
Part II: Analysis Unit ............................Part II: 4-1  
Appendix......................................................... A-1  
Teledyne Analytical Instruments  
iii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 5000B  
iv  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
OPERATING INSTRUCTIONS  
Model 5000B  
Photometric Analyzer  
Part I: Control Unit  
NEMA 4 Bulkhead Mount  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 5000B Photometric Analyzer  
Table of Contents  
1 Introduction  
1.1 Overview........................................................................ 1-1  
1.2 Typical Applications ....................................................... 1-1  
1.3 Main Features of the Analyzer ....................................... 1-5  
1.4 Operator Interface .......................................................... 1-6  
1.4.1 Up/Down Switch .................................................... 1-6  
1.4.2 Escape/Enter Switch.............................................. 1-6  
1.5 Control Section Interface Panel...................................... 1-8  
2 Installation  
2.1 Unpacking the Control Unit/Analysis Unit ...................... 2-1  
2.2 Electrical Connections ................................................... 2-1  
2.3 Testing the System ........................................................ 2-9  
3 Operation  
3.1 Introduction .................................................................... 3-1  
3.2 Using the Data Entry and Function Buttons ................... 3-1  
3.3 The System Function ..................................................... 3-4  
3.3.1 Setting up an Auto-Cal........................................... 3-4  
3.3.2 Password Protection .............................................. 3-6  
3.3.2.1 Entering the Password ................................... 3-6  
3.3.2.2 Installing or Changing the Password ............. 3-7  
3.3.3 Logging Out ........................................................... 3-8  
3.3.4 System Self-Diagnostic Test .................................. 3-9  
3.3.5 The Model Screen ................................................. 3-10  
3.3.6 Checking Linearity with Algorithm ......................... 3-10  
3.3.7 Digital Flter Setup .................................................. 3-11  
3.3.8 Filter or Solenoid Setup ......................................... 3-12  
3.3.9 Hold/Track Setup ................................................... 3-13  
3.3.10 Calibration/Hold Timer Setup ................................ 3-13  
3.3.11 Analog 4 to 20 mA Output Calibration.................... 3-14  
ii: Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Part I: Control Unit  
3.3.12 Model ..................................................................... 3-15  
3.3.13 Show Negative ...................................................... 3-15  
3.4 The Zero and Span Functions ....................................... 3-16  
3.4.1 Zero Cal ................................................................. 3-16  
3.4.1.1 Auto Mode Zeroing ........................................ 3-16  
3.4.1.2 Manual Mode Zeroing.................................... 3-17  
3.4.1.3 Cell Failure .................................................... 3-18  
3.4.2 Span Cal................................................................ 3-19  
3.4.2.1 Auto Mode Spanning ..................................... 3-19  
3.4.2.2 Manual Mode Spanning................................. 3-19  
3.4.3 Offset Function....................................................... 3-20  
3.5 The Alarms Function...................................................... 3-22  
3.6 The Range Select Function ........................................... 3-24  
3.6.1 Manual (Select/Define Range) Screen .................. 3-24  
3.6.2 Auto Screen ........................................................... 3-25  
3.6.3 Precautions............................................................ 3-26  
3.7 The Analyze Function .................................................... 3-28  
3.8 Programming ................................................................. 3-28  
3.8.1 The Set Range Screen .......................................... 3.29  
3.8.2 The Curve Algorithm Screen ................................. 3-31  
3.8.2.1 Manual Mode Linearization ........................... 3-31  
3.8.2.2 Auto Mode Linearization ................................ 3-31  
4 Maintenance  
4.1 Fuse Replacement......................................................... 4-1  
4.2 System Self Diagnostic Test........................................... 4-2  
4.3 Major Internal Components............................................ 4-3  
Teledyne Analytical Instruments  
Part I: iii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 5000B Photometric Analyzer  
iv: Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Photometric Analyzer  
Part I: Control Unit  
Introduction  
1.1 Overview  
The Teledyne Analytical Instruments Model 5000B Control Unit,  
together with a 5000B Analysis Unit, is versatile microprocessor-based  
instrument.  
Part I, of this manual covers the Model 5000B General Purpose NEMA  
4 Bulkhead Mount Control Unit. (The Analysis Unit is covered in Part II of  
this manual.) The Control Unit is for indoor/outdoor use in a nonhazardous  
environment only. The Control or Analysis Unit in the folded optics design  
can accomodate a variety of hazardous environments with either full explo-  
sion proof housings or Z, X, Y or Cenelec Purging functions.  
1.2 Typical Applications  
A few typical applications of the Model 5000B are:  
WATERMONITORING  
Background  
Acids,including:  
Acetic  
Typical Range  
0–4000 PPM  
Formic  
Sulfuric  
Acetaldehyde  
Air  
0–1000 PPM  
0–2%  
Alcohols,including:  
Butanol  
0–400 ppm  
Part I: 1-1  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1 Introduction  
Model 5000B  
Ethanol  
Isopropanol  
Methanol  
Alkanes,including:  
Heptane  
0–500 ppm  
Hexane  
Ammonia  
0–1000 ppm and up  
0–500 ppm  
Aromatics,including:  
Benzene  
Cumene  
Toluene  
Xylene  
ChlorinatedHydrocarbonsincluding:  
CarbonTetrachloride  
EthylChloride  
0–200 ppm  
EthyleneDichloride  
MethylChloride  
Perchloroethylene  
PropyleneDichloride  
Trichloroethylene  
Vinyl Chloride  
Chloroprene  
Chloropicrin  
Deuterium Oxide  
Epichlorohydrin  
Ethylene Glycol  
Freons  
0–200 ppm  
0–200 ppm  
0–200 ppm  
0–2000 ppm  
0–500 ppm  
0–500 ppm  
0–500 ppm  
Gasoline  
1-2: Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Photometric Analyzer  
Part I: Control Unit  
HydrogenFluoride 0–10%  
Hydroperoxides  
Kerosene  
0–5%  
0–500 ppm  
0–1000 ppm  
0–1000 ppm  
Ketones  
MethylAcetate  
MethylMethacrylate 0–1000 ppm  
Oils  
0–1%  
Olefins  
0–500 ppm  
0–300 ppm  
0–300 ppm  
0–1000 ppm  
Pentane  
α-Picolene  
Phenol  
Polyols  
0–500 ppm  
Propylene Glycol  
Propylene Oxide  
Sulfinol  
0–500 ppm  
0–200 ppm  
0–15%  
SulfurDioxide  
Vinyl Acetate  
0–1000 ppm  
0–2%  
NOTE:Range may be higher or lower per application.  
OTHER NIR ABSORBERS (that can be measured)  
(consult factory)  
Acetic Acid  
Alcohols  
Amines  
Aromatics  
Butadiene  
Carbonyls  
Chloroprene  
Esters  
Part I: 1-3  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1 Introduction  
Model 5000B  
Hydroxyl Value  
Hydrocarbons  
Hydrogen Chloride  
Hydrogen Fluoride  
Hydroxyl Value  
Ketones  
Olefins  
Oximes  
Epoxides  
Methylene  
Ketones  
1-4: Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Photometric Analyzer  
Part I: Control Unit  
1.3 Main Features of the Analyzer  
The Model 5000B Photometric Analyzer is sophisticated yet simple to  
use. The main features of the analyzer include:  
A 2-line alphanumeric display screen, driven by microprocessor  
electronics, that continuously prompts and informs the operator.  
High resolution, accurate readings of concentration from low  
ppm levels through to 100%. Large, bright, meter readout.  
Versatile analysis over a wide range of applications.  
Microprocessorbasedelectronics:8-bitCMOSmicroprocessor  
with 32 kB RAM and 128 kB ROM.  
Three user definable output ranges (from 0-1 ppm through  
0-100 %) allow best match to users process and equipment.  
Calibration range for convenient zeroing or spanning.  
Auto Ranging allows analyzer to automatically select the proper  
preset range for a given measurement. Manual override allows  
the user to lock onto a specific range of interest.  
Two adjustable concentration alarms and a system failure alarm.  
Extensive self-diagnostic testing, at startup and on demand, with  
continuouspower-supplymonitoring.  
RS-232 serial digital port for use with a computer or other digital  
communicationdevice.  
Analog outputs for concentration and range identification.  
(0-1 V dc standard, and isolated 4–20 mA dc)  
Superioraccuracy.  
Internalcalibration-ManualorAutomatic(optional).  
Part I: 1-5  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1 Introduction  
Model 5000B  
1.4 Operator Interface  
All controls and displays on the standard 5000B are accessible from  
outside the housing. The instrument has two simple operator controls. The  
operator has constant feedback from the instrument through an alphanumeric  
display, and a digital LED meter. The displays and controls are described  
briefly here and in greater detail in chapter 3. See Figure 1-1.  
1.4.1 UP/DOWN Switch  
The UP/DOWN switch is used to select between any subfunctions  
displayed on the VFD screen such as in the main menue, the system menue,  
the Alarm menue, etc. When modifiable values are displayed on the VFD,  
the UP/DOWN switch can be used to increment or decrement the values.  
1.4.2 ESCAPE/ENTER Switch  
The ESCAPE/ENTER switch is used to input the data, to enter a  
function, or to exit a function displayed in the alphanumeric display:  
Escape  
Moves VFD display back to the previous screen in a  
series. If none remains, returns to Analyze mode  
screen.  
Enter  
moving on  
Within a menue: the funtion selected is entered  
to the next screen in a series.  
With Value selected: Enters the value into the  
data. Advances cursor on VFD to the  
analyzeras  
nextoperation.  
In the Analyze mode: it calls the main menue.  
Functions  
number  
called out by the main menue:  
-System This function is a menu that calls a  
offunctions  
that regulate the analyzer  
operation.  
-Span  
-Zero  
This function spans the instrument.  
This function zeros the instrument.  
-Alarms This functions sets the alarm  
preferences.  
1-6: Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Photometric Analyzer  
Part I: Control Unit  
-Range Thisfunctionselectswhetheranalyzeris  
autoranging or locked on one range.  
-Standby Places the analyzer in a sleep mode.  
WARNING:  
The power cable must be disconnected to fully remove  
power from the instrument.  
Figure 1-1: Model 5000B Controls, Indicators, and Connectors  
Digital Meter Display: The meter display is a Light Emitting Diode  
LED device that produces large, bright, 7-segment numbers that are legible  
in any lighting. It is accurate across all analysis ranges. The 5000B models  
produce continuous readout from 0-10,000 ppm and then switch to  
continuous percent readout from 1-100 %.  
Part I: 1-7  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1 Introduction  
Model 5000B  
Alphanumeric Interface Screen: The backlit VFD screen is an easy-  
to-use interface between operator and analyzer. It displays values, options,  
and messages for immediate feedback to the operator.  
1.5 Control Section Interface Panel  
The Control Section interface panel, shown in Figure 1-2, contains the  
electrical terminal blocks for external inputs and outputs. The input/output  
functions are described briefly here and in detail in the Installation chapter of  
thismanual.  
Power Connection  
Analog Outputs  
AC power source, 115VAC, 50/60 Hz  
0-1 V dc concentration and 0-1 V dc  
range ID. Isolated 4-20 mA dc and 4-20  
mA dc range ID.  
Alarm Connections  
RS-232 Port  
2 concentration alarms and 1 system  
alarm.  
Serialdigitalconcentrationsignaloutput  
and control input.  
Remote Bench  
Providesallelectricalinterconnecttothe  
AnalysisSection.  
Remote Span/Zero  
Digital inputs allow external control of  
analyzercalibration.  
Calibration Contact To notify external equipment that  
instrument is being calibrated and  
readings are not monitoring sample.  
Range ID Contacts  
Network I/O  
Four separate, dedicated, range relay  
contacts.  
Serialdigitalcommunicationsforlocal  
network access. For future expansion.  
Not implemented at this printing.  
Note: If you require highly accurate Auto-Cal timing, use external  
Auto-Cal control where possible. The internal clock in the  
Model 5000B is accurate to 2-3 %. Accordingly, internally  
scheduled calibrations can vary 2-3 % per day.  
1-8: Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Photometric Analyzer  
Part I: Control Unit  
Figure 1-2: Model 5000B Interface Panel  
Part I: 1-9  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1 Introduction  
Model 5000B  
1-10: Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Photometric Analyzer  
Part I: Control Unit  
Installation  
Installation of Model 5000B Analyzers includes:  
1. Unpacking, mounting, and interconnecting the Control/Analysis  
Section  
2. Making gas connections to the system  
3. Making electrical connections to the system  
4. Testing the system.  
This chapter covers installation of the Control Section. (Installation of  
the Analysis Section is covered in Part II of this manual.)  
2.1 Unpacking the Control/Analysis Unit  
The analyzer is shipped with all the materials you need to install and  
prepare the system for operation. Carefully unpack the Control/Analysis  
Unit and inspect it for damage. Immediately report any damage to the ship-  
ping agent. Figure 2-2: Required Front Door Clearance  
Allow clearance for the door to open in a 90-degree arc of radius 15.5  
inches. See Figure 2-2.  
Figure 2-2: Required Front Door  
Clearance  
2.2 Electrical Connections  
Figure 2-3 shows the Control/Analysis Unit interface panel. Connec-  
tions for power, communications, and both digital and analog signal outputs  
are described in the following paragraphs. Wire size and maximum length  
data appear in the Drawings at the back of this manual.  
Part I: 2-1  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1 Introduction  
Model 5000B  
Figure 2-3: Interface Panel of the Model 6600 Control Section  
For safe connections, ensure that no uninsulated wire extends  
outside of the terminal blocks. Stripped wire ends must insert completely  
into terminal blocks. No uninsulated wiring should come in contact with  
fingers, tools or clothing during normal operation.  
Primary Input Power: The power supply in the Model 6600 will  
accept a 115 Vac, 50/60 Hz power source. See Figure 2-4 for detailed  
connections.  
DANGER: Power is applied to the instrument's circuitry as  
long as the instrument is connected to the power  
source. The standby function switches power on or  
off to the displays and outputs only.  
2-2: Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Photometric Analyzer  
Part I: Control Unit  
Figure 2-4: Primary Input Power Connections  
115VAC,  
Fuse Installation: The fuse holders accept 5 x 20 mm, 4.0 A, T type  
(slow blow) fuses. Fuses are not installed at the factory. Be sure to install  
the proper fuse as part of installation (See Fuse Replacement in chapter 4,  
maintenance.)  
Analog Outputs: There are eight DC output signal connectors on  
the ANALOG OUTPUTS terminal block. There are two connectors per  
output with the polarity noted. See Figure 2-5.  
The outputs are:  
0–1 V dc % of Range: Voltage rises linearly with increasing sample con-  
centration, from 0 V at 0% to 1 V at 100%. (Full  
scale = 100% programmed range.)  
0–1 V dc Range ID:  
0.25 V = Range 1, 0.5 V = Range 2, 0.75 V =  
Range 3.  
4–20 mA dc % Range: Currentincreaseslinearlywithincreasingsample  
concentration, from 4 mA at 0% to 20 mA at full  
scale 100%. (Full scale = 100% of programmed  
range.)  
Part I: 2-3  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1 Introduction  
Model 5000B  
4–20 mA dc Range ID: 8 mA = Range 1, 12 mA = Range 2, 16 mA =  
Range 3.  
Figure 2-5: Analog Output Connections  
Examples:  
The analog output signal has a voltage which depends on the sample  
concentration AND the currently activated analysis range. To relate the  
signal output to the actual concentration, it is necessary to know what range  
the instrument is currently on, especially when the analyzer is in the  
autorangingmode.  
The signaloutput for concentration is linear over currently selected  
analysis range. For example, if the analyzer is set on a range that was  
defined as 0-10 %, then the output would be as shown in Table 2-1.  
Table 2-1: Analog Concentration Output-Examples  
Concentration Voltage Signal  
Current Signal  
Output (mA dc)  
%
Output (V dc)  
0
0.0  
4.0  
5.6  
7.2  
1
2
0.1  
0.2  
2-4: Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Photometric Analyzer  
Part I: Control Unit  
3
4
0.3  
0.4  
0.5  
0.6  
0.7  
0.8  
0.9  
1.0  
8.8  
10.4  
12.0  
13.6  
15.2  
16.8  
18.4  
20.0  
5
6
7
8
9
10  
To provide an indication of the range, a second pair of analog output  
terminals are used. They generate a steady preset voltage (or current when  
using the current outputs) to represent a particular range. Table 2-2 gives the  
range ID output for each analysis range.  
Table 2-2: Analog Range ID Output - Example  
Range  
Voltage (V)  
0.25  
Current (mA)  
Range 1  
Range 2  
Range 3  
8
0.50  
12  
16  
0.75  
Alarm Relays:  
There are three alarm-circuit connectors on the alarm relays block  
(under RELAY OUTPUTS) for making connections to internal alarm relay  
contacts. Each provides a set of Form C contacts for each type of alarm.  
Each has both normally open and normally closed contact connections. The  
contact connections are indicated by diagrams on the rear panel. They are  
capable of switching up to 3 ampers at 250 V AC into a resistive load  
(Figure2-6).  
Part I: 2-5  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1 Introduction  
Model 5000B  
Normally closed  
Normally open  
Moving contact  
Normally open  
Moving contact  
Figure  
2-6: Types of Relay Contacts  
The connectors are:  
Threshold Alarm 1: • Can be configured as high (actuates when  
concentration is above threshold), or low  
(actuates when concentration is below thresh old).  
• Can be configured as fail-safe or non-fail-safe.  
• Can be configured as latching or nonlatching.  
• Can be configured out (defeated).  
Threshold Alarm 2: • Can be configured as high (actuates when concen-  
tration is above threshold), or low (actuates when  
concentrationisbelowthreshold).  
• Can be configured as fail-safe or non-fail-safe.  
• Can be configured as latching or nonlatching.  
• Can be configured out (defeated).  
SystemAlarm:  
Actuates when DC power supplied to circuits is  
unacceptable in one or more parameters. Permanently  
configured as fail-safe and latching. Cannot be de-  
feated. Actuates if self test fails.  
To reset a System Alarm during installation, discon-  
nect power to the instrument and then reconnect it  
Further detail can be found in chapter 3, section 3-5.  
Digital Remote Cal Inputs  
Remote Zero and Span Inputs: The REMOTE SPAN and RE-  
MOTE ZERO inputs are on the DIGITAL INPUT terminal block. They  
2-6: Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Photometric Analyzer  
Part I: Control Unit  
accept 0 V (OFF) or 24 V dc (ON) for remote control of calibration (See  
Remote Calibration Protocol below.)  
Zero: Floating input. 5 to 24 V input across the + and – terminals  
puts the analyzer into the ZERO mode. Either side may be  
grounded at the source of the signal. 0 to 1 volt across the  
terminals allows ZERO mode to terminate when done. A  
synchronous signal must open and close the external zero  
valve appropriately. See Remote Probe Connector at end of  
section 3.3. (With the -C option, the internal valves automati-  
callyoperatesynchronously).  
Span:  
Floating input. 5 to 24 V input across the + and – terminals  
puts the analyzer into the SPAN mode. Either side may be  
grounded at the source of the signal. 0 to 1 volt across the  
terminals allows SPAN mode to terminate when done. A  
synchronous signal must open and close the external span  
valve appropriately. See Remote Probe Connector at end of  
section 3.3. (With the -C option, the internal valves automati-  
callyoperatesynchronously.)  
Cal Contact: This relay contact is closed while analyzer is spanning  
and/or zeroing. (See Remote Calibration Protocol below.)  
Remote Calibration Protocol: To properly time the Digital Remote  
Cal Inputs to the Model 5000B Analyzer, the customer's controller must  
monitor the Cal Relay Contact.  
When the contact is OPEN, the analyzer is analyzing, the Remote Cal  
Inputs are being polled, and a zero or span command can be sent.  
When the contact is CLOSED, the analyzer is already calibrating. It  
will ignore your request to calibrate, and it will not remember that request.  
Once a zero or span command is sent, and acknowledged (contact  
closes), release it. If the command is continued until after the zero or span is  
complete, the calibration will repeat and the Cal Relay Contact (CRC) will  
closeagain.  
Forexample:  
1) Test the CRC. When the CRC is open, Send a zero command  
until the CRC closes (The CRC will quickly close.)  
2) When the CRC closes, remove the zero command.  
3) When CRC opens again, send a span command until the CRC  
closes. (The CRC will quickly close.)  
4) When the CRC closes, remove the span command.  
Part I: 2-7  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1 Introduction  
Model 5000B  
When CRC opens again, zero and span are done, and the sample is  
being analyzed.  
Note: The Remote Bench terminal strip (section 3.6 Part III) provides  
signals to ensure that the zero and span gas valves will be  
controlled synchronously.  
Range ID Relays: Four dedicated RANGE ID CONTACT relays .  
The first four ranges are assigned to relays in ascending order—Range 1 is  
assigned to RANGE 1 ID, Range 2 is assigned to RANGE 2 ID, Range 3  
is assigned to RANGE 3 ID, and Range 4 is assigned to RANGE 4 ID.  
Network I/O: A serial digital input/output for local network protocol.  
At this printing, this port is not yet functional. It is to be used in future  
versions of the instrument.  
RS-232 Port: The digital signal output is a standard RS-232 serial  
communications port used to connect the analyzer to a computer, terminal, or  
other digital device. The pinouts are listed in Table 2-3.  
Table 2-3: RS-232 Signals  
RS-232 Sig RS-232 Pin Purpose  
DCD  
RD  
1
2
3
4
5
6
7
8
9
Data Carrier Detect  
Received Data  
TransmittedData  
Data Terminal Ready  
Common  
TD  
DTR  
COM  
DSR  
RTS  
CTS  
RI  
Data Set Ready  
Request to Send  
Clear to Send  
RingIndicator  
The data sent is status information, in digital form, updated every two  
seconds. Status is reported in the following order:  
The concentration in percent  
The range is use (HI< MED< LO)  
The span of the range 0-100%, etc.  
Which alarm - if any - are disabled (AL-x DISABLED)  
2-8: Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Photometric Analyzer  
Part I: Control Unit  
Which alarms - if any - are tripped (AL-x ON)  
Eachstatusoutputisfollowedbyacarriagereturnandlinefeed.  
ThreeinputfunctionsusingRS-232havebeenimplementedtodate. They  
aredescribedinTable2-4.  
Table 2-4: Commands via RS-232 Input  
Command  
as<enter>  
az<enter>  
st<enter>  
Description  
Immediatelystartsanautospan.  
Immediatelystartsanautozero.  
Togglinginput. Stops/Startsanystatusmessageoutput  
fromtheRS-232,Untilst<enter>issentagain.  
TheRS-232protocolallowssomeflexibilityinitsimplementation.Table2-  
5listscertainRS-232valuesthatarerequiredbythe5000B.  
Table 2-5: Required RS-232 Options  
Parameter  
Baud  
Setting  
2400  
Byte  
Parity  
8 bits  
none  
Stop Bits  
MessageInterval  
1
2 seconds  
Remote Bench and Solenoid Valves: The 5000B is a single-chassis  
instrument. However, the REMOTE BENCH and SOLENOID RETURN  
connectors are provided on the interface PCB. The Remote Bench is wired  
at the factory as well as any optional solenoid valves included in the system.  
2.3 Testing the System  
AfterTheControl/AnalysisUnitisbothinstalledandinterconnected,and  
thesystemgasorliquidstreamandelectricalconnectionsarecomplete,the  
systemisreadytotest.Beforepluggingtheunitintoitspowersources:  
Checktheintegrityandaccuracyofthefluidconnections.Makesure  
therearenoleaks.  
Checktheintegrityandaccuracyofallelectricalconnections.Make  
suretherearenoexposedconductors  
Part I: 2-9  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1 Introduction  
Model 5000B  
Checkthatsamplepressuretypicallybetween0and30psig,  
according to the requirements of your process.  
Power up the system, and test it by performing the following operation:  
1. Repeat the Self-Diagnostic Test.  
2. Zero the instrument.  
3. Span the instrument.  
Consult calibration recommendation in the addendum for your  
particular application whether liquid or gas phase monitoring.  
For steps 2 and 3, refer to part II for calibration.  
2-10: Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Photometric Analyzer  
Operation /Control Unit 3  
Operation  
3.1 Introduction  
Although the Model 5000B is usually programmed to your application  
at the factory, it can be further configured at the operator level, or even,  
cautiously, reprogrammed. Depending on the specifics of the application,  
this might include all or a subset of the following procedures:  
Settingsystemparameters:  
Establish a security password, if desired, requiring Operator  
to log in (secure in safe file for referrence).  
Establish and start an automatic calibration cycle, if desired.  
RoutineOperation:  
Calibratetheinstrument.  
Choose autoranging or select a fixed range of analysis.  
Set alarm setpoints, and modes of alarm operation (latching,  
fail-safe,etc).  
Program/Reprogramtheanalyzer:  
Define new applications.  
Linearize your ranges.  
If you choose not to use password protection, the default password is  
automatically displayed on the password screen when you start up, and you  
simply press Enter for access to all functions of the analyzer.  
3.2 Using the Data Entry and Function  
Buttons  
Data Entry Buttons: The < > buttons select options from the menu  
currently being displayed on the VFD screen. The selected option blinks.  
When the selected option includes a modifiable item, the DÑ arrow  
buttons can be used to increment or decrement that modifiable item.  
Teledyne Analytical Instruments  
Part I 3-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3 Operation/Control Unit  
Model 5000B  
The Enter button is used to accept any new entries on the VFD screen.  
The Escape button is used to abort any new entries on the VFD screen that  
are not yet accepted by use of the Enter button.  
Figure3-1showsthehierarchyoffunctionsavailabletotheoperatorviathe  
functionbuttons.Thesixfunctionsofthebuttonsontheanalyzerare:  
Analyze.Thisisthenormaloperatingmode.Theanalyzermonitors  
theconcentrationofthesample,displaysthepercentorparts-per-  
millionoftargetfluidorcontamination,andwarnsofanyalarm  
conditions.  
System.Thesystemfunctionconsistsofninesubfunctions.  
Fouroftheseareforordinarysetupandoperation:  
SetupanAuto-Cal  
AssignPasswords  
Logouttosecuresystem  
InitiateaSelf-Test  
Threeofthesubfunctionsdoauxiliarytasks:  
Checkingmodelandsoftwareversion  
Adjustelectronicfilterofthesignal  
Displaymoresubfunctions  
Displaynegativereadings  
Twooftheseareforprogramming/reprogrammingtheanalyzer:  
Definefluidapplicationsandranges(Refertoprogramming  
section, or contact factory.)  
Use the Curve Algorithm to linearize output. (Refer to  
programmingsection, orcontactfactory.)  
Zero. Used to set up a zero calibration.  
Span. Used to set up a span calibration.  
Alarms. Used to set the alarm setpoints and determine whether  
each alarm will be active or defeated, HI or LO acting, latching,  
and/orfail-safe.  
Range. Used to set up three analysis ranges that can be switched  
automatically with autoranging or used as individual fixed  
ranges.  
Any function can be selected at any time by pressing the appropriate  
button (unless password restrictions apply). The order as presented in this  
manual is appropriate for an initial setup.  
Eachofthesefunctionsisdescribedingreaterdetailinthefollowingproce-  
dures.TheVFDscreentextthataccompanieseachoperationisreproduced,at  
3-2 Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Photometric Analyzer  
Operation /Control Unit 3  
System  
SetDigital  
Filter  
Dig_filt  
Self-Testin  
Progress  
Self-Test  
Results  
SELF-TEST  
Change  
Yes/No  
Change  
Password  
Verify  
Password  
Enter  
Password  
Enter  
PWD  
SecureSystem  
LOGOUT  
setupnotallowed  
MORE  
Span/Zerostatus  
and<>setup  
Span/Zerotiming  
andon/off  
AUTOCAL  
Enter  
Span/Zero  
FILSOL  
Enter  
SolenoidorFilter  
TRACK  
Settrackor  
holdoutput  
or  
HOLD  
Setcal.holdand  
sampleholdtimer  
CAL-HOLD  
TIMER  
Enter  
MORE  
Verifydata  
Points  
Enter  
Displaygasuse  
andrange  
Select  
Enter  
Selectrange  
Selectrange  
ALGORITHM  
APPLICATION  
Enter  
Verify/Setup  
Input/Output  
Auto/Manual  
linearCal.  
Define  
Application/Range  
Enter  
EnterSpan  
Enter  
gasvalue  
OUTPUT:  
4or20MA  
Setcurrent  
output  
Enter  
Enter  
MORE  
Display  
Model/Version  
MODEL  
ShowNegative  
Reading  
SHOW.NEG  
Figure 3-1: Hierarchy of System Functions and Subfunctions  
Teledyne Analytical Instruments  
Part I 3-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3 Operation/Control Unit  
Model 5000B  
theappropriatepointintheprocedure,inaMonospacedtypestyle.Push-button  
namesareprintedinObliquetype.  
3.3 The System Function  
ThesubfuctionsoftheSystemfunctionaredescribedbelow.Specific  
procedures for their use follow the descriptions:  
Dig_Filt: Adjust how much digital filtering should be on the  
signal  
SELF-TEST: Performs a self-diagnostic test to check the  
integrity of the power supplies, outputs, detector signal and  
preamplifier.  
PWD: Login security system for accessing to the setup functions.  
LOGOUT: Prevents an unauthorized tampering with analyzer  
settings.  
AUTOCAL: Set the automatic calibrated timer schedule for  
Zero and Span cycling.  
FILSOL: Select Span/Zero flag (filter) or Span/Zero solenoid  
valveforcalibrationmethod.  
TRACK: Set the system reading to be held or followed by the  
concentrationfluidorfilterduringcalibration.  
CAL-HOLD-TIMER: Set the timing for calibration holding  
and timing for the sample reading after return to analyze mode.  
ALGORITHM: Linearize the output for nonlinear characteristic.  
APPLICATION: Used to define the analysis ranges and  
application(fluidused).  
MODEL: Displays model number and software version.  
OUTPUT: 4-20 MA: Adjust 4 and 20 mA output.  
SHOW_NEG: Whether to display negative readings or not;  
affects analog output too. No negative readings is the default.  
3.3.1 Setting up an AUTO-CAL  
When proper automatic valving is connected, the Analyzer can cycle  
itself through a sequence of steps that automatically zero and span the instru-  
ment.  
3-4 Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Photometric Analyzer  
Operation /Control Unit 3  
Note: Before setting up an AUTO-CAL, be sure you understand the Zero  
and Span functions as described in section 4.4, and follow the  
precautions given there.  
Note: If you require highly accurate AUTO-CAL timing, use external  
AUTO-CAL control where possible. The internal clock in the  
Model 5000BF is accurate to 2-3 %. Accordingly, internally  
scheduled calibrations can vary 2-3 % per day.  
To setup an Auto–Cal cycle:  
Choose System from the Function buttons. TheVFD will display five  
subfunctions.  
DIG_FILT SELF—TEST  
PWD LOGOUT MORE  
Select MORE and press the Enter Key  
AUTOCAL  
CAL-HOLD-TIMER  
FILSOL  
MORE  
HOLD  
Use < > arrows to blink AUTO—CAL, and press Enter. A new screen  
for ZERO/SPAN set appears.  
ZERO in Ød Øh off  
SPAN in Ød Øh off  
Press < > arrows to blink ZERO (or SPAN), then press Enter again.  
(You won’t be able to set OFF to ON if a zero interval is entered.) A Span  
Every ... (or Zero Every ...) screen appears.  
Zero schedule: OFF  
Day: Ød Hour: Øh  
Use DÑ arrows to set an interval value, then use < > arrows to move to  
the start-time value. Use DÑ arrows to set a start-time value.  
To turn ON the SPAN and/or ZERO cycles (to activate AUTO–CAL):  
Press System again, choose AUTO–CAL, and press Enter again. When the  
ZERO/SPAN values screen appears, use the < > arrows to blink the ZERO  
(or SPAN) and press Enter to go to the next screen. Use < > to select OFF/  
ON field. Use DÑ arrows to set the OFF/ON field to ON. You can now turn  
these fields ON because there is a nonzero span interval defined.  
If instrument is turned off, the next time the instrument is powered, the  
instrument will automatically perform a calibration cycle after 3 minutes of  
entering the sample mode if AUTOCAL functions were on prior to shut  
down.  
Teledyne Analytical Instruments  
Part I 3-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3 Operation/Control Unit  
Model 5000B  
3.3.2 PasswordProtection  
Beforeauniquepasswordisassigned,thesystemassignsTAIbydefault.  
Thispasswordwillbedisplayedautomatically.Theoperatorjustpressesthe  
Enterkeytobeallowedtotalaccesstotheinstrument’sfeatures.  
Ifapasswordisassigned,thensettingthefollowingsystemparameterscan  
bedoneonlyafterthepasswordisentered:alarmsetpoints,assigninganew  
password,range/applicationselections,andcurvealgorithmlinearization.  
(APPLICATIONandALGORITHMarecoveredintheprogrammingsection.)  
However,theinstrumentcanstillbeusedforanalysisorforinitiatingaself-test  
withoutenteringthepassword.Todefeatsecuritythepasswordmustbe  
changedbacktoTAI.  
NOTE: If you use password security, it is advisable to keep a copy of the  
password in a separate, safe location.  
3.3.2.1  
Entering the Password  
To install a new password or change a previously installed password,  
you must key in and ENTER the old password first. If the default password  
is in effect, pressing the ENTER button will enter the default TAI password  
for you.  
Press System to enter the System mode.  
DIG_FILT AUTO—CAL  
PWD LOGOUT MORE  
Use the < > arrow keys to scroll the blinking over to PWD, and press  
Enter to select the password function. Either the default TAI password or  
AAA place holders for an existing password will appear on screen depend-  
ing on whether or not a password has been previously installed.  
Enter password:  
T A I  
or  
Enter password:  
A A A  
The screen prompts you to enter the current password. If you are not  
using password protection, press Enter to accept TAI as the default pass-  
word. If a password has been previously installed, enter the password using  
the < > arrow keys to scroll back and forth between letters, and the DÑ  
3-6 Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Photometric Analyzer  
Operation /Control Unit 3  
arrow keys to change the letters to the proper password. Press Enter to enter  
thepassword.  
Inafewseconds,youwillbegiventheopportunitytochangethispass-  
word or keep it and go on.  
Change Password?  
<ENT>=Yes <ESC>=No  
Press Escape to move on, or proceed as in Changing the Password,  
below.  
3.3.2.2  
Installing or Changing the Password  
If you want to install a password, or change an existing password,  
proceed as above in Entering the Password. When you are given the oppor-  
tunity to change the password:  
Change Password?  
<ENT>=Yes <ESC>=No  
Press Enter to change the password (either the default TAI or the  
previously assigned password), or press Escape to keep the existing pass-  
word and move on.  
If you chose Enter to change the password, the password assignment  
screenappears.  
Select new password  
T A I  
Enter the password using the < > arrow keys to move back and forth  
between the existing password letters, and the DÑ arrow keys to change the  
letters to the new password. The full set of 94 characters available for pass-  
word use are shown in the table below.  
Characters Available for Password Definition:  
A
K
U
_
i
s
}
)
B
L
V
`
j
t
®
*
4
>
C
M
W
a
k
u
!
+
5
D
N
X
b
l
v
"
'
E
O
Y
c
m
w
#
F
P
Z
d
n
x
$
.
G
Q
[
H
R
¥
f
p
z
&
0
:
I
S
]
g
q
{
'
1
;
J
T
^
h
r
|
(
e
o
y
%
/
-
7
2
<
3
=
6
@
8
9
?
Teledyne Analytical Instruments  
Part I 3-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3 Operation/Control Unit  
Model 5000B  
Whenyouhavefinishedtypingthenewpassword,pressEnter.Averifica-  
tionscreenappears.Thescreenwillpromptyoutoretypeyourpasswordfor  
verification.  
Enter PWD To Verify:  
A A A  
Use the arrow keys to retype your password and press Enter when  
finished. Your password will be stored in the microprocessor and the system  
will immediately switch to the Analyze screen, and you now have access to  
allinstrumentfunctions.  
If all alarms are defeated, the Analyze screen appears as:  
1.95 ppm SO2  
nR1: Ø — 1Ø Anlz  
If an alarm is tripped, the second line will change to show which alarm  
itis:  
1.95 ppm SO2  
AL—1  
NOTE:  
If you log off the system using the LOGOUT function in the  
system menu, you will now be required to re-enter the password to  
gain access to Alarm, and Range functions.  
3.3.3 Logging Out  
The LOGOUT function provides a convenient means of leaving the  
analyzer in a password protected mode without having to shut the instrument  
off. By entering LOGOUT, you effectively log off the instrument leaving  
the system protected against use until the password is reentered. To log out,  
press the System button to enter the System function.  
DIG_FILT SELF-TEST  
PWD LOGOUT MORE  
Use the < > arrow keys to position the blinking over the LOGOUT  
function, and press Enter to Log out. The screen will display the message:  
Protected until  
password entered  
3-8 Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Photometric Analyzer  
Operation /Control Unit 3  
3.3.4 System Self-Diagnostic Test  
TheModel5000BFhasabuilt-inself-diagnostictestingroutine.Pre-  
programmed signals are sent through the power supply, output board,  
preamp board and sensor circuit. The return signal is analyzed, and at the  
end of the test the status of each function is displayed on the screen, either as  
OK or as a number between 1 and 1024. (See System Self Diagnostic Test in  
chapter 4 for number code.) If any of the functions fails, the System Alarm is  
tripped.  
Note: The sensor will always show failed unless Zero fluid is present in  
the sampling cell at the time of the SELF-TEST.  
The self diagnostics are run automatically by the analyzer whenever the  
instrument is turned on, but the test can also be run by the operator at will.  
To initiate a self diagnostic test during operation:  
Press the System button to start the System function.  
DIG_FILT SELF-TEST  
PWD LOGOUT MORE  
Use the < > arrow keys again to move the blinking to the SELF–TEST  
and press Enter. The screen will follow the running of the diagnostic.  
RUNNING DIAGNOSTIC  
Testing Preamp — Cell  
When the testing is complete, the results are displayed.  
Power: OK Analog: OK  
Cell: 2 Preamp: 3  
The module is functioning properly if it is followed by OK. A number  
indicates a problem in a specific area of the instrument. Refer to Chapter 5  
Maintenance and Troubleshooting for number-code information. The results  
screen alternates for a time with:  
Press Any Key  
To Continue...  
Then the analyzer returns to the initial System screen.  
Teledyne Analytical Instruments  
Part I 3-9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3 Operation/Control Unit  
Model 5000B  
3.3.5 The Model Screen  
Move the < > arrow key to MORE and press Enter. With MODEL  
blinking, press Enter. The screen displays the manufacturer, model, and  
softwareversioninformation.  
3.3.6 CheckingLinearitywithALGORITHM  
FromtheSystemFunctionscreen,selectALGORITHM,andpressEnter.  
sel rng to set algo:  
—> Ø1 Ø2 Ø3 <—  
Use the < > keys to select the range: 01, 02, or 03. Then press Enter.  
Fluid Use: SO2  
Range:  
Ø — 10%  
Press Enter again.  
Algorithm setup:  
VERIFY SET UP  
Select and Enter VERIFY to check whether the linearization has been  
accomplishedsatisfactorily.  
Dpt INPUT OUTPUT  
Ø Ø.ØØ  
Ø.ØØ  
The leftmost digit (under Dpt) is the number of the data point being  
monitored. Use the DÑ keys to select the successive points.  
The INPUT value is the input to the linearizer. It is the simulated output  
of the analyzer. You do not need to actually flow fluid.  
The OUTPUT value is the output of the linearizer. It should be the  
ACTUAL concentration of the span fluid being simulated.  
If the OUTPUT value shown is not correct, the linearization must be  
corrected. Press ESCAPE to return to the previous screen. Select and Enter  
SET UP to Calibration Mode screen. (set-up will not work without a PC  
being connected to the analyzer)  
Select algorithm  
mode : AUTO  
There are two ways to linearize: AUTO and MANUAL: The auto  
mode requires as many calibration fluides as there will be correction points  
alongthecurve.Theuserdecidesonthenumberofpoints,basedonthepreci-  
sionrequired.  
3-10 Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Photometric Analyzer  
Operation /Control Unit 3  
Themanualmodeonlyrequiresenteringthevaluesforeachcorrection  
point into the microprocessor via the front panel buttons. Again, the number  
of points required is determined by the user.  
3.3.7 Digital Filter Setup  
The 5000BF has the option of decreasing or increasing the amount  
filtering on the signal. This feature enhances the basic filtering done by the  
analog circuits by setting the amount of digital filtering effected by the  
microprocessing. To access the digital filter setup, you must:  
1.  
Press the System key to start the System function  
DIG_FILT SELF-TEST  
PWD LOGOUT MORE  
2.  
3.  
DIG_FILT will flash, press the ENTER key,  
Weight of digital  
Filter:  
9
The number on the second row will flash and can be set by  
using the Up or Down arrow keys.  
The settings go from zero, no digital filtering, to 10, maximum digital  
filtering. The default setting is 8 and that should suffice for most applica-  
tions. In some applications where speeding the response time with some  
trade off in noise is of value, the operator could decrease the number of the  
digital filter. In applications where the signal is noisy, the operator could  
switch to a higher number; the response time is slowed down though.  
90% response time on the different settings to a step input is shown  
below. This response time does not include the contributions of the bench  
sampling system and the preamplifier near the detector.  
Setting 90% Response time  
(seconds)  
0
1
2
3
4
5
4.5  
4.5  
5.0  
5.0  
5.5  
7.0  
Teledyne Analytical Instruments  
Part I 3-11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3 Operation/Control Unit  
Model 5000B  
6
9.0  
7
14.0  
25.0  
46.0  
90.0  
8
9
10  
Atasettingofzero”,theresponsetimeispurelysetbytheelectronicsto  
4.5seconds. Thenumbersabovecanandwillchangedependingonapplication  
andtheymerelyservetoillustratetheeffectofthedigitalfilter.  
3.3.8 FilterorSolenoidSetup  
The5000BFcanbespannedorzeroedbycalibrationfluidsorbyoptical  
filters. Thepropercalibrationmethodshouldbesetatthefactory. Toaccess  
theFilterorSolenoidFlags,youmust:  
1. PresstheSystemkeytostarttheSystemfunction:  
DIG_FILT SELF-TEST  
PWD LOGOUT MORE  
2. Using the Right or Left arrow keys, select MORE and press Enter.  
The second System screen appears:  
AUTOCAL FILSOL TRACK  
CAL-HOLD-TIMER MORE  
3. Select FILSOL using the Right or Left arrow keys and press Enter to  
start the method of calibration function.  
Set fil/sol for cal  
Span: FIL Zero: SOL  
There are two flag options: zero and/or span flags are choosen at time  
of purchase, one for Zero calibration and the other for Span located in the  
Detector housing. To move between the Zero and the Span flags, use the  
Right or Left arrow keys. FIL means that a filter will do this particular  
calibration. SOL means that the signal to activate a fluid solenoid is enabled.  
To toggle between the SOL and FIL options, use the Up and Down arrow  
keys.  
Theconnectionstodrivethefilterandthesolenoidarefoundonastrip  
terminallocatedontheinterfaceboard. Theconnectionsaredescribedin  
section 5-6 of the maintenance section of the manual.  
3-12 Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Photometric Analyzer  
Operation /Control Unit 3  
3.3.9  
Hold/Track Setup  
The5000BFhasabilitytodisabletheanalogoutputsandfreezethedisplay  
whileundergoingascheduledorremotecalibration. The5000BFwilltrack  
changesintheconcentrationifcalibrationisstartedthroughthefront  
panel. Tosetupthisfeature,theoperatormust:  
1.  
PresstheSystemkeytostarttheSystemfunction:  
DIG_FILT SELF-TEST  
PWD LOGOUT MORE  
2.  
Using the Right or Left arrow keys, select MORE and press  
Enter. The Second System screen appears:  
AUTOCAL FILSOL TRACK  
CAL-HOLDER-TIMER MORE  
or  
AUTOCAL FILSOL HOLD  
CAL-HOLD-TIMER MORE  
3.  
The option on the right of the first row can be set to TRACK  
or HOLD by using the UP or Down keys. By selecting the TRACK option,  
the analog outputs are enabled and with the display will track the concentra-  
tion changes while the instrument is undergoing scheduled or remote calibra-  
tion (either zero or span). By selecting the HOLD option, the analog outputs  
and display are disabled and will not track the concentration changes while  
the instrument is undergoing scheduled or remote calibration (either zero or  
span). In the HOLD option, the analog outputs and display will freeze on  
the last reading before entering calibration.  
The analog outputs are both 0 to 1 volt outputs and both 4 to 20 mA  
outputs.  
3.3.10 Calibration/Hold Timer Setup  
This Calibration Timer lets the operator adjust the time thee instrument  
purges the calibration fluid prior to actually start the calibration computa-  
tions. The Sample timer lets the operator adjust the time the instrument  
purgessamplefluidafterfinishingacalibrationbeforeitletstheanalogoutputs  
anddisplaytrackthechangeinconcentration.  
Teledyne Analytical Instruments  
Part I 3-13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3 Operation/Control Unit  
Model 5000B  
ThisfunctionandtheTRACK/HOLDfeaturewillpreventfalsealarms  
while performing remote or autoscheduled calibrations. These functions are  
not applicable if the calibration is initiated through the front panel. To enter  
the Calibration/Hold Timer function, you must:  
1.  
Press the System key to start the System function:  
DIG_FILT SELF-TEST  
PWD LOGOUT MORE  
2.  
Using the Right or Left arrow keys, select MORE and press  
Enter: The Second System screen appears:  
AUTOCAL FILSOL TRACK  
CAL-HOLD-TIMER MORE  
or  
AUTOCAL FILSOL HOLD  
CAL-HOLD-TIMER MORE  
3.  
Select with the Right or Left keys CAL-HOLD-TIMER, and  
press the Enter key to access this function menu:  
Calbrt hold: 3 min  
Sample hold: 1 min  
The calibration hold time is set on the first row, while the sample hold  
time is set on the second row. To select one or the other, use the Right or  
Left keys. To modify the time of either timer, use the Up or Down keys.  
The time is in the minutes.  
3.3.11 Analog 4 to 20 mA Output Calibration  
This function will let the operator calibrate the 4 to 20 mA analog  
output to match the display reading. A DMM configure as a DC ammeter is  
needed. The DMM should be connected across the output terminals of the 4  
to 20 mA output to monitor the output current. To enter the 4 to 20 mA  
output adjust function, you must:  
1.  
Press the System key to start the System function:  
DIG_FILT SELF-TEST  
PWD LOGOUT MORE  
2.  
Using the Right or Left arrow keys, select MORE and press  
Enter. The second System screen appears:  
AUTOCAL FILSOL TRACK  
CAL-HOLD-TIMER MORE  
3-14 Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Photometric Analyzer  
Operation /Control Unit 3  
or  
AUTOCALFILSOLHOLD  
CAL-HOLD-TIMER MORE  
3.  
Using the Right or the Left arrow keys, select MORE and  
press Enter. The third System screen appears:  
ALGORITHM APPLICATION  
MODEL OUTPUT: 4 MA  
or  
ALGORITHM APPLICATION  
MODEL OUTPUT: 20 MA  
OUTPUT: 4 MA and OUTPUT: 20 MA can be toggled by moving on  
that field and pressing the Up or Down key. 4 mA output should be calibrat-  
ed first and 20 mA output afterwards.  
4.  
Select OUTPUT: 4 MA and press the Enter key  
Use UP/DOWN arrow to  
Adjust 4 ma:  
250  
The number on the second row is the setpoint of the 4 mA output. It is  
analogous to a potentiometer wiper. The number can be set anywhere from  
0 to 500. The default is 250, in the middle. At the default setting, the output  
should be very close to 4 mA. If not, slowly adjust the number using the Up  
or the Down keys until DMM reads 4.00 mA. Press the Enter key when  
done.  
5.  
Now select OUTPUT: 20 MA and press the Enter key. A  
screen similar to the one above will appear and the DMM should read close  
to 20 mA. If not, slowly adjust the number using the Up or Down key until  
DMM reads 20.0 mA. Press the Enter key when done.  
The range of adjustment is approximately +/- 10% of scale (+/- 1.6 ma).  
Since the 4 to 20 mA output is tied to the 0 to 1 volt output, this function can  
be used to calibrate the 0 to 1 volt output, if the 4 to 20 mA output is not  
used. By using a digital Volt meter on the 0-1 Volt output.  
3.3.12 Model  
This selection in the System menu flashes for a few seconds the model  
number and the software version installed in this instrument.  
Teledyne Analytical Instruments  
Part I 3-15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3 Operation/Control Unit  
Model 5000B  
3.3.13 Show Negative  
Theanalyzerdefaultstonottoshownegativereadingsontheanalyzemode  
only. ThisaffectstheanalogoutputstoobypressingtheUPorDOWNkey,the  
analyzercanbesettodisplaynegativereadings,ontheSHOW_NEGfieldof  
thesystemmenu.  
3.4 The Zero and Span Functions  
TheModel5000BFcanhaveasmanyasthreeanalysisrangesplusa  
special calibration range (Cal Range). Calibrating any one of the ranges will  
automaticallycalibratetheotherranges.  
CAUTION: Always allow 4-5 hours warm-up time before calibrating, if  
your analyzer has been disconnected from its power source.  
This does not apply if the analyzer was plugged in but was in  
STANDBY.  
The analyzer is calibrated using zero, and span fluides.  
Note: Shut off the fluid pressure before connecting it to the analyzer, and  
be sure to limit pressure to 40 psig or less when turning it back on.  
Readjust the fluid pressure into the analyzer until the flowrate through  
the Sample Cell settles between 50 to 200 cc/min (approximately 0.1 to 0.4  
SCFH).  
Note: Always keep the calibration fluid flow as close to the flowrate of the  
sample fluid as possible  
3.4.1 Zero Cal  
The Zero function on the menu is used to enter the zero calibration  
function. Zero calibration can be performed in either the automatic or manual  
mode.  
Make sure the zero fluid is flowing to the instrument. If you get a  
CELL CANNOT BE BALANCED message while zeroing skip to section  
4.4.1.3.  
3.4.1.1  
Auto Mode Zeroing  
Observe the precautions in sections 4.4 and 4.4.1, above. Press Zero to  
enter the zero function mode. The screen allows you to select whether the  
3-16 Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Photometric Analyzer  
Operation /Control Unit 3  
zerocalibrationistobeperformedautomaticallyormanually. UsetheDÑ  
arrow keys to toggle between AUTO and MAN zero settling. Stop when  
AUTO appears, blinking, on the display.  
Select zero  
mode: AUTO  
Press Enter to begin zeroing.  
####.## %  
SO2  
Slope=#.### C—Zero  
The beginning zero level is shown in the upper left corner of the dis-  
play. As the zero reading settles, the screen displays and updates information  
on Slope= in percent/second (unless the Slope starts within the acceptable  
zero range and does not need to settle further). The system first does a coarse  
zero, shown in the lower right corner of the screen as C—Zero, for 3 min, and  
then does a fine zero, and displays F—Zero, for 3 min.  
Then, and whenever Slope is less than 0.01 for at least 3 min, instead of  
Slope you will see a countdown: 9 Left, 8 Left, and so fourth. These are  
software steps in the zeroing process that the system must complete, AFTER  
settling, before it can go back to Analyze. Software zero is indicated by S–  
Zero in the lower right corner.  
####.## % SO2  
4 Left=#.### S—Zero  
The zeroing process will automatically conclude when the output is  
within the acceptable range for a good zero. Then the analyzer automatically  
returns to the Analyze mode.  
3.4.1.2  
Manual Mode Zeroing  
Press Zero to enter the Zero function. The screen that appears allows  
you to select between automatic or manual zero calibration. Use the DÑ  
keys to toggle between AUTO and MAN zero settling. Stop when MANU-  
AL appears, blinking, on the display.  
Select zero  
mode: MANUAL  
Press Enter to begin the zero calibration. After a few seconds the first  
of three zeroing screens appears. The number in the upper left hand corner is  
the first-stage zero offset. The microprocessor samples the output at a prede-  
terminedrate.  
Teledyne Analytical Instruments  
Part I 3-17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3 Operation/Control Unit  
Model 5000B  
####.##  
% SO2  
Zero adj:2048 C—Zero  
The analyzer goes through C–Zero, F–Zero, and S–Zero. During C–  
Zero and F–Zero, use the DÑ keys to adjust displayed Zero adj: value as  
close as possible to zero. Then, press Enter.  
S–Zero starts. During S–Zero, the Microcontroller takes control as in  
Auto Mode Zeroing, above. It calculates the differences between successive  
samplings and displays the rate of change as Slope= a value in parts per  
million per second (ppm/s).  
####.##  
%
SO2  
Slope=#.### S—Zero  
Generally, you have a good zero when Slope is less than 0.05 ppm/s for  
about 30 seconds.  
Once zero settling completes, the information is stored in the analyzer’s  
memory, and the instrument automatically returns to the Analyze mode.  
3.4.1.3 Cell Failure  
Detector failure in the 5000BF is usually associated with inability to  
zero the instrument with a reasonable voltage differential between the refer-  
ence and measure voltages. If this should ever happen, the 5000BF system  
alarm trips, and the LCD displays a failure message.  
Detector cannot be balanced  
Check your zero fluid  
Beforeopticalbalancing:  
a. Check your zero fluid to make sure it is within specifications.  
b. Check for leaks downstream from the Sample Cell, where con-  
tamination may be leaking into the system.  
c. Check flowmeter to ensure that the flow is no more than 200  
SCCM  
d. Checktemperaturecontrollerboard.  
e. Checkfluidtemperature.  
f. Check the Sample Cell for dirty windows.  
If none of the above, proceed to perform an optical balance as described  
in section 5.  
3-18 Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Photometric Analyzer  
Operation /Control Unit 3  
3.4.2 SpanCal  
TheSpanfunctiononthemenuisusedtospancalibratetheanalyzer.Span  
calibrationcanbeperformedineithertheautomaticormanualmode.  
Makesurethespanfluidisflowingtotheinstrument.  
3.4.2.1  
AutoModeSpanning  
Observeallprecautionsinsections3.4and3.4.2, above. PressSpanto  
enter the span function. The screen that appears allows you to select whether  
the span calibration is to be performed automatically or manually. Use the  
DÑ arrow keys to toggle between AUTO and MAN span settling. Stop  
when AUTO appears, blinking, on the display.  
Select span  
mode: AUTO  
Press Enter to move to the next screen.  
Span Val: 2Ø.ØØ %  
<ENT> To begin span  
Use the < > arrow keys to toggle between the span concentration value  
and the units field (%/ppm). Use the DÑ arrow keys change the value and/  
or the units, as necessary. When you have set the concentration of the span  
fluid you are using, press Enter to begin the Span calibration.  
####.##%  
Slope=#.###  
SO2  
Span  
The beginning span value is shown in the upper left corner of the  
display. As the span reading settles, the screen displays and updates informa-  
tion on Slope. Spanning automatically ends when the span output corre-  
sponds, within tolerance, to the value of the span fluid concentration. Then  
the instrument automatically returns to the analyze mode.  
3.4.2.2  
Manual Mode Spanning  
Press Span to start the Span function. The screen that appears allows  
you to select whether the span calibration is to be performed automatically or  
manually.  
Select span  
mode: MANUAL  
Teledyne Analytical Instruments  
Part I 3-19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3 Operation/Control Unit  
Model 5000B  
UsetheDÑkeystotogglebetweenAUTOandMANspansettling. Stop  
whenMANappears,blinking,onthedisplay.PressEntertomovetothenext  
screen.  
Span Val: 2Ø.ØØ %  
<ENT> To begin span  
Use the < > arrow keys to toggle between the span concentration value  
and the units field (%/ppm). Use the DÑ arrow keys change the value and/  
or the units, as necessary. When you have set the concentration of the span  
fluid you are using, press Enter to begin the Span calibration.  
Press Enter to enter the span value into the system and begin the span  
calibration.  
Once the span has begun, the microprocessor samples the output at a  
predetermined rate. It calculates the difference between successive samplings  
and displays this difference as Slope on the screen. It takes several seconds  
for the first Slope value to display. Slope indicates rate of change of the Span  
reading. It is a sensitive indicator of stability.  
####.##%  
SO2  
Slope=#.###  
Span  
When the Span value displayed on the screen is sufficiently stable,  
press Enter. (Generally, when the Span reading changes by 1 % or less of  
the range being calibrated for a period of ten minutes it is sufficiently stable.)  
Once Enter is pressed, the Span reading changes to the correct value. The  
instrument then automaticallyenters the Analyze function.  
3.4.3 OffsetFunction  
This software when installed in the 5000BF instruments provides a way to  
enter an offset on the zero operation of the analyser. This is useful when the  
instrument is zeroed in some inert fluid such as Nitrogen or Argon, but the  
background fluid of the process is different. If the background fluid of the  
process is different than the zero calibration fluid being used, the reading will  
have an offset that will be constant throughout its working range. Thus, the  
need to provide an offset when the instrument is being zeroed.  
How to access the offset function:  
To access this function, the instrument zero mode must be entered by  
pushing the Zero key on the front panel of the control unit. The VFD display  
will show the following menu selection:  
3-20 Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Photometric Analyzer  
Operation /Control Unit 3  
Select zero  
mode: AUTO  
or  
Select zero  
mode: MAN  
Select whether you want the instrument to do an automatic or manual zero. If  
you do an automatic zero, the instrument does the zero by itself. If you do a  
manual zero you must manually enter inputs to the instrument to accomplish  
the zero, see in the corresponding section of the manual on how these two  
functionsdiffer.  
When the Enter key is pressed, the following menu will appear:  
Zero off: 0.0 ppm  
<ENT> to begin Zero  
The offset value can be modified by using the Up/Down keys. Next section  
shows how to select this value. Suffice to say that whatever value you enter,  
will be automatically added to the reading. Thus, if you entered -0.1 ppm, at  
the end of the zero the display will show -0.1 ppm.  
Once the Enter key is pressed the instrument enters the zero mode. If you  
chose AUTO zero mode, the instrument will do the work of bringing the  
reading back to zero plus the offset value that was entered. If you chose  
MANual zero mode, then you must enter input to the instrument as  
explained in the corresponding section of the manual but with one  
difference: instead of bringing the display to read zero, you must make the  
display read zero plus the value entered as offset.  
How the offset value is selected:  
To find out what the offset value should be, the intended zero calibration  
fluid and the a mix of the process background fluid must be procured. This  
of course assumes that the zero fluid and the process background fluid are  
very different and that an offset will occur.  
1. Let the intended zero calibration fluid flow through the 5000BF sample  
cell (this assumes that you have started up you system as recommended by  
the manual or technical personnel) and do a zero on the instrument. Leave  
the offset set to zero value.  
2. At the end of the zero function, make sure the analyser reads zero.  
3. Flow a zero level of the process background fluid mix through the  
5000BF sample cell on the Analyse mode. Wait for the reading to become  
stable. Write the reading down. Change the sign of the reading: This is the  
Teledyne Analytical Instruments  
Part I 3-21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3 Operation/Control Unit  
Model 5000B  
offsettobeentered.  
4. Do a manual run to check. Reintroduce the zero calibration fluid. Start a  
zero on the analyser but this time enter the offset value.  
5. At the end of the zero function, check that the instrument reads the entered  
offset.  
6. Reintroduce the process background fluid mix to the 5000BF sample cell  
in the Analyse mode. It should read close to zero once the reading is stable  
(+/- 1% error of full scale).  
Spanning the 5000BF:  
Since the instrument might be spanned with background fluid the same as  
the zero calibration fluid, the span value to be entered should be the span  
concentration plus the offset value (if the offset value has a minus sign then  
algebraicallyitbecomesasubtraction).  
3.5 The Alarms Function  
The Model 5000BF is equipped with two fully adjustable set points  
concentration with two alarms and a system failure alarm relay. Each alarm  
relay has a set of form “C" contacts rated for 3 amperes resistive load at 250  
V ac. See Figure in Chapter 2, Installation and/or the Interconnection Dia-  
gram included at the back of this manual for relay terminal connections.  
The system failure alarm has a fixed configuration described in chapter  
2Installation.  
The concentration alarms can be configured from the front panel as  
either high or low alarms by the operator. The alarm modes can be set as  
latching or non-latching, and either failsafe or non-failsafe, or, they can be  
defeated altogether. The setpoints for the alarms are also established using  
thisfunction.  
Decide how your alarms should be configured. The choice will depend  
upon your process. Consider the following four points:  
1. Which if any of the alarms are to be high alarms and which if any  
are to be low alarms?  
Setting an alarm as HIGH triggers the alarm when the  
contaminant concentration rises above the setpoint. Setting an  
alarm as LOW triggers the alarm when the contaminant  
concentration falls below the setpoint.  
Decide whether you want the alarms to be set as:  
3-22 Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Photometric Analyzer  
Operation /Control Unit 3  
Bothhigh(highandhigh-high)alarms,or  
One high and one low alarm, or  
Both low (low and low-low) alarms.  
2. Are either or both of the alarms to be configured as failsafe?  
In failsafe mode, the alarm relay de-energizes in an alarm  
condition. For non-failsafe operation, the relay is energized in an  
alarm condition. You can set either or both of the concentration  
alarms to operate in failsafe or non-failsafe mode.  
3. Are either of the alarms to be latching?  
In latching mode, once the alarm or alarms trigger, they will  
remain in the alarm mode even if process conditions revert back  
to non-alarm conditions. This mode requires an alarm to be  
recognized before it can be reset. In the non-latching mode, the  
alarm status will terminate when process conditions revert to non-  
alarmconditions.  
4. Are either of the alarms to be defeated?  
The defeat alarm mode is incorporated into the alarm circuit so  
that maintenance can be performed under conditions which  
would normally activate the alarms.  
The defeat function can also be used to reset a latched alarm.  
(See procedures, below.)  
If you are using password protection, you will need to enter your  
password to access the alarm functions. Follow the instructions in section  
3.3.3 to enter your password. Once you have clearance to proceed, enter the  
Alarm function.  
Press the Alarm button on the front panel to enter the Alarm function.  
Use the DÑ keys to choose between % or ppm units. Then press Enter  
to move to the next screen.  
AL1: 1ØØØ ppm HI  
Dft:N Fs:N Ltch:N  
Five parameters can be changed on this screen:  
Value of the alarm setpoint, AL1: ####  
Out-of-range direction, HI or LO  
Defeated? Dft:Y/N (Yes/No)  
Failsafe? Fs:Y/N (Yes/No)  
Teledyne Analytical Instruments  
Part I 3-23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3 Operation/Control Unit  
Model 5000B  
Latching?Ltch:Y/N(Yes/No).  
Todefinethesetpoint,usethe<>arrowkeystomovetheblinking  
over to AL1: ####. Then use the DÑ arrow keys to change the  
number.Holdingdownthekeyspeedsuptheincrementingor  
decrementing.  
To set the other parameters use the < > arrow keys to move the  
blinking over to the desired parameter. Then use the DÑ arrow  
keys to change the parameter.  
Once the parameters for alarm 1 have been set, press Alarms  
again, and repeat this procedure for alarm 2 (AL2).  
To reset a latched alarm, go to Dft– and then press either D two  
times or Ñ two times. (Toggle it to Y and then back to N.)  
–OR –  
Go to Ltch– and then press either D two times or Ñ two times.  
(Toggle it to N and back to Y.)  
3.6 The Range Select Function  
The Range function allows you to manually select the concentration  
range of analysis (MANUAL), or to select automatic range switching  
(AUTO).  
In the MANUAL screen, you are further allowed to define the high and  
low (concentration) limits of each Range, and select a single, fixed range to  
run.  
CAUTION: If this is a linearized application, the new range must be  
within the limits previously programmed using the System  
function, if linearization is to apply throughout the range.  
Furthermore, if the limits are too small a part (approx 10 %  
or less) of the originally linearized range, the linearization  
willbecompromised.  
3.6.1 Manual (Select/Define Range) Screen  
The Manual range-switching mode allows you to select a single, fixed  
analysis range. It then allows you to redefine the upper and lower limits, for  
the range.  
Press Range key to start the Range function.  
3-24 Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Photometric Analyzer  
Operation /Control Unit 3  
Select range  
mode: MANUAL  
If above screen displays, use the DÑ arrow keys to Select MANUAL,  
and press Enter.  
Select range to run  
—> Ø1 Ø2 Ø3 CAL<—  
Use the < > keys to select the range: 01, 02, 03, or CAL. Then press  
Enter.  
Fluid use: SO2  
Range: Ø — 10 %  
Use the < > keys to toggle between the Range: low-end field and the  
Range: high-end field. Use the DÑ keys to change the values of the fields.  
Press Escape to return to the previous screen to select or define another  
range.  
Press Enter to return the to the Analyze function.  
3.6.2 Auto Screen  
Autoranging will automatically set to the application that has at least  
two ranges setup with the same fluides.  
In the autoranging mode, the microprocessor automatically responds to  
concentration changes by switching ranges for optimum readout sensitivity.  
If the upper limit of the operating range is reached, the instrument automati-  
cally shifts to the next higher range. If the concentration falls to below 85%  
of full scale of the next lower range, the instrument switches to the lower  
range. A corresponding shift in the DC concentration output, and in the  
range ID outputs, will be noticed.  
The autoranging feature can be overridden so that analog output stays  
on a fixed range regardless of the contaminant concentration detected. If the  
concentration exceeds the upper limit of the range, the DC output will  
saturate at 1 V dc (20 mA at the current output).  
However, the digital readout and the RS-232 output of the concentra-  
tion are unaffected by the fixed range. They continue to read beyond the full-  
scalesettinguntilamplifiersaturationisreached. Belowamplifiersaturation,  
the overrange readings are accurate UNLESS the application uses lineariza-  
tion over the selected range.  
Teledyne Analytical Instruments  
Part I 3-25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3 Operation/Control Unit  
Model 5000B  
TheconcentrationrangescanberedefinedusingtheRangefunction  
Manual screen, and the application fluides can be redefined using the System  
function, if they are not already defined as necessary.  
CAUTION: Redefining applications or ranges might require relineariza-  
tionand/orrecalibration.  
To setup automatic ranging:  
Press Range key to start the Range function.  
Select range  
mode : AUTO  
If above screen displays, use the DÑ arrow keys to Select AUTO, and  
press Enter.  
Press Escape to return to the previous Analyze Function.  
3.6.3 Precautions  
The Model 5000BF allows a great deal of flexibility in choosing ranges  
for automatic range switching. However, there are some pitfalls that are to be  
avoided.  
Ranges that work well together are:  
Ranges that have the same lower limits but upper limits that differ  
by approximately an order of magnitude  
Ranges whose upper limits coincide with the lower limits of the  
next higher range  
Ranges where there is a gap between the upper limit of the range  
and the lower limit of the next higher range.  
Range schemes that are to be avoided include:  
Ranges that overlap  
Ranges whose limits are entirely within the span of an adjoining  
range.  
Ranges where the zero is suppressed, is 1-10, 1-100, etc,  
however, 80-100, 90-100 is ok where the zero fluid is actually  
100% concentration and the calibration is inverted.  
3-26 Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Photometric Analyzer  
Operation /Control Unit 3  
Figure3-2illustratestheseschemesgraphically.  
0.01 0.1 80  
0
90  
100  
Figure 3-2: Examples of Autoranging Schemes  
Teledyne Analytical Instruments  
Part I 3-27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3 Operation/Control Unit  
Model 5000B  
3.7 The Analyze Function  
Normally,allofthefunctionsautomaticallyswitchbacktotheAnalyze  
function when they have completed their assigned operations. Pressing the  
Escape button in many cases also switches the analyzer back to the Analyze  
function. Alternatively, you can press the Analyze button at any time to  
return to analyzing your sample.  
The Analyze function screen shows the impurity concentration and the  
application fluides in the first line, and the range in the second line. In the  
lower right corner, the abbreviation Anlz indicates that the analyzer is in the  
Analyze mode. If there is an * before the Anlz, it indicates that the range is  
linearized.  
1.95 ppm  
SO2  
R1:Ø —10 *Anlz  
If the concentration detected is overrange, the first line of the display  
blinkscontinuously.  
3.8 Programming  
CAUTION: The programming functions of the Set Range and Curve  
Algorithm screens are configured at the factory to the users  
application specification. These functions should only be  
reprogrammed by trained, qualified personnel.  
To program, you must:  
1. Enter the password, if you are using the analyzer’s password  
protectioncapability.  
2. Connect a computer or computer terminal capable of sending an  
RS-232 signal to the analyzer RS-232 connector. (See chapter 2  
Installation for details). Send the rp command to the analyzer.  
3. Press the System button to start the System function.  
DIG_FILT SELF-TEST  
PWD LOGOUT MORE  
Use the < > arrow keys to blink MORE, then press Enter.  
3-28 Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Photometric Analyzer  
Operation /Control Unit 3  
AUTOCAL FILSOLL HOLD  
CAL-HOLD-TIMER MORE  
Select MORE and press ENTER one more time  
ALGORITHM  
APPLICATION  
MORE OUTPUT: 4MA  
Now you will be able to select the APPLICATION and ALGO-  
RITHM set-up functions.  
3.8.1 The Set Range Screen  
The Set Range screen allows reprogramming of the three analysis  
ranges and the calibration range (background fluid, low end of range, high  
end of range, and % or ppm units). Original programming is usually done at  
the factory according to the customer’s application. It must be done through  
the RS-232 port using a computer running a terminal emulation program.  
Note: It is important to distinguish between this System programming  
subfunction and the Range button function, which is an operator  
control. The Set Range Screen of the System function allows the  
user to DEFINE the upper and lower limits of a range AND the  
application of the range. The Range button function only allows the  
user to select or define the limits, or to select the application, but  
not to define the application.  
Normally the Model 5000BF is factory set to default to manual range  
selection, unless it is ordered as a single-application multiple-range unit (in  
which case it defaults to autoranging). In either case, autoranging or manual  
range selection can be programmed by the user.  
In the autoranging mode, the microprocessor automatically responds to  
concentration changes by switching ranges for optimum readout sensitivity.  
If the upper limit of the operating range is reached, the instrument automati-  
cally shifts to the next higher range. If the concentration falls to below 85%  
of full scale of the next lower range, the instrument switches to the lower  
range. A corresponding shift in the DC concentration output, and in the  
range ID outputs, will be noticed.  
The autoranging feature can be overridden so that analog output stays  
on a fixed range regardless of the contaminant concentration detected. If the  
concentration exceeds the upper limit of the range, the DC output will  
saturate at 1 V dc (20 mA at the current output).  
However, the digital readout and the RS-232 output of the concentra-  
tion are unaffected by the fixed range. They continue to read beyond the full-  
Teledyne Analytical Instruments  
Part I 3-29  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3 Operation/Control Unit  
Model 5000B  
scalesettinguntilamplifiersaturationisreached.Belowamplifiersaturation,the  
overrangereadingsareaccurateUNLESStheapplicationuseslinearizationover  
theselectedrange.  
Toprogramtheranges,youmustfirstperformthefourstepsindicatedat  
thebeginningofsection3.8Programming.Youwillthenbeinthesecond  
System menu screen.  
ALGORITHM APPLICATION  
MORE  
OUTPUT: 4MA  
Use the < > arrow keys again to move the blinking to APPLICATION  
and press Enter.  
Sel rng to set appl:  
—> Ø1 Ø2 Ø3 CAL <—  
Use the DÑ arrow keys to increment/decrement the range number to  
01, 02, 03, or CAL, and press Enter.  
Fluid Name  
**********  
FR:Ø TO:1Ø %  
Use the < > arrow keys to move to Fluid Name, FR: (from—lower end  
of range), TO: (to—upper end of range), and PPM or %.  
Use the DÑ arrow keys to increment the respective parameters as  
desired. Press Enter to accept the values and return to Analyze mode. (See  
note below.) Repeat for each range you want to set.  
Note: The ranges must be increasing from low to high, for example, if  
Range 1 is set to 0–10 % and Range 2 is set to 0–100 %, then  
Range 3 cannot be set to 0–50 % since that makes Range 3 lower  
than Range 2.  
Ranges, alarms, and spans are always set in either percent or ppm units,  
as selected by the operator, even though all concentration-data outputs  
change from ppm to percent when the concentration is above 9999 ppm.  
Note: When performing analysis on a fixed range, if the concentration  
rises above the upper limit as established by the operator for that  
particular range, the output saturates at 1 V dc (or 20 mA). How-  
ever, the digital readout and the RS-232 output continue to read  
regardless of the analog output range.  
To end the session, send:  
st<enter>  
st<enter>  
3-30 Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Photometric Analyzer  
Operation /Control Unit 3  
totheanalyzerfromthecomputer.  
3.8.2 TheCurveAlgorithmScreen  
TheCurveAlgorithmisalinearizationmethod.Itprovidesfrom1to9  
intermediate points between the ZERO and SPAN values, which can be  
normalizedduringcalibration, toensureastraight-lineinput/outputtransfer  
function through the analyzer.  
Each range is linearized individually, as necessary, since each range  
willusuallyhaveatotallydifferentlinearizationrequirement.  
To linearize the ranges, you must first perform the four steps indicated  
at the beginning of section 3.8 Programming. You will then be in the second  
System menu screen.  
3.8.2.1 Manual Mode Linearization  
To linearize manually, you must have previous knowledge of the  
nonlinear characteristics of your fluides. You enter the value of the differen-  
tial between the actual concentration and the apparent concentration (analyz-  
er output). TAI has tabular data of this type for a large number of fluides,  
which it makes available to customers on request. See Appendix for ordering  
information. To enter data:  
From the System Functions Screen—  
1. Use < > to select ALGORITHM , and Enter.  
2. Select and Enter SETUP.  
3. Enter MANUAL from the Calibration Mode Select screen.  
Dpt INPUT OUTPUT  
Ø Ø.ØØ  
Ø.ØØ  
The data entry screen resembles the verify screen, but the fluid values  
can be modified and the data-point number cannot. Use the < > keys to  
toggle between the INPUT and OUTPUT fields. Use the DÑ keys to set the  
value for the lowest concentration into the first point. Then press Enter.  
After each point is entered, the data-point number increments to the  
next point. Moving from the lowest to the highest concentration, use the DÑ  
keys to set the proper values at each point.  
Dpt INPUT OUTPUT  
0 Ø.ØØ  
Ø.ØØ  
Teledyne Analytical Instruments  
Part I 3-31  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3 Operation/Control Unit  
Model 5000B  
Repeattheaboveprocedureforeachofthedatapointsyouaresetting(up  
toninepoints:0-8).Setthepointsinunitincrements.Donotskipnumbers.The  
linearizerwillautomaticallyadjustforthenumberofpointsentered.  
Whenyouaredone, PressESCAPE. Themessage, Completed. Waitfor  
calculation,appearsbriefly,andthenthemainSystemscreenreturns.  
Toendthesession,send:  
st<enter>  
st<enter>  
to the analyzer from the computer.  
3.8.2.2 Auto Mode Linearization  
To linearize in the Auto Mode, you must have on hand a separate  
calibration fluid for each of the data points you are going use in your linear-  
ization. First, the analyzer is zeroed and spanned as usual. Then, each special  
calibration fluid, for each of the intermediate calibration points, is flowed, in  
turn, through the sensor. As each fluid flows, the differential value for that  
intermediate calibration point is entered from the front panel of the analyzer.  
Note: The span fluid used to span the analyzer must be >90% of the  
range being analyzed.  
Before starting linearization, perform a standard calibration. See section  
4.4. To enter data:  
From the System Functions screen—  
1. Use < > to select ALGORITHM , and Enter.  
2. Select and Enter SETUP.  
3. Enter AUTO from the Calibration Mode Select screen.  
The Auto Linearize Mode data entry screen appears.  
19.5 ppm SO2  
Input(Ø) :20.00  
5. Use the DÑ keys to set the proper value of calibration fluid, and  
Enter. Repeat this step for each cal-point number as it appears in  
the Input (x) parentheses.  
6. Repeat step 5 for each of the special calibration fluides, from the  
lowest to the highest concentrations. Press Escape when done.  
To end the session, send:  
st<enter>  
st<enter>  
to the analyzer from the computer.  
3-32 Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Part I: Control Unit/Analysis Unit  
Maintenance 3  
Maintenance  
Aside from normal cleaning and checking for leaks at the gas connections,  
routine maintenance is limited to replacing filter elements and fuses, and  
recalibration.  
WARNING: SEE WARNINGS ON THE TITLE PAGE OF THIS  
MANUAL.  
4.1 Fuse Replacement  
The 5000B requires two 5 x 20 mm, 6.3 A, F type (Fast Blow) fuses.  
The fuses are located inside the main housing on the Electrical  
Connector Panel, as shown in Figure 4-3. To replace a fuse:  
1. Disconnect the Unit from its power source.  
2. Place a small screwdriver in the notch in the fuse holder  
cap, push in, and rotate 1/4 turn. The cap will pop out a  
few millimeters. Pull out the fuse cap and fuse, as  
shown in Figure 4-1  
Part I: 4-1  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3 Maintenance  
Model 5000B Photometric Analyzer  
Figure 4-1: Removing Fuse Block Cap and Fuse from Housing  
2. Replace fuse by reversing process in step 1.  
4.2 System Self Diagnostic Test  
NOTE: Always run self diagnostic with the intended zero fluid.  
1. Press the System button to enter the system mode.  
2. Use the < > arrow keys to move to More, and press Enter.  
3. Use the < > arrow keys to move to Self-Test, and press Enter.  
The following failure codes apply:  
Table 4-1: Self Test Failure Codes  
Power  
0
1
2
3
OK  
5 V Failure  
15 V Failures  
Both Failed  
Analog  
0
1
2
3
OK  
DAC A (0–1 V Concentration)  
DAC B (0–1 V Range ID)  
Both Failed  
Teledyne Analytical Instruments  
4-2: Part I  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Part I: Control Unit/Analysis Unit  
Maintenance 3  
Preamp  
0
>0  
OK  
means that some of the ten gains of the  
amplifier have a large offset. The number is a  
code that tells which of the ten gtains have the  
largeoffset.  
Cell(detector)  
0
OK  
>0  
means that the analyzer failed to  
balance the measuring and the reference  
signals, the larger the number the farther off  
they were.4.3 MajorInternalComponents  
The major components in the Control Unit are shown in Figure 4-3.  
Figure 4-3: Control Unit Major Internal Components  
Part I: 4-3  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3 Maintenance  
Model 5000B Photometric Analyzer  
WARNING: HAZARDOUS VOLTAGES EXIST ON CERTAIN  
COMPONENTS INTERNALLY WHICH MAY PERSIST  
FOR A TIME EVEN AFTER THE POWER IS TURNED  
OFF AND DISCONNECTED.  
The 5000B Control Units contain the following major components:  
Power Supply  
Motherboard (with Microprocessor, RS-232 chip, and  
PreamplifierPCB)  
Front Panel Display Board and Displays—  
5 digit LED meter  
2 line, 20 character, alphanumeric, VFD display  
See the drawings in the Drawings section or addendum in back of  
this manual for details.  
Teledyne Analytical Instruments  
4-4: Part I  
Download from Www.Somanuals.com. All Manuals Search And Download.  
OPERATING INSTRUCTIONS  
Model 5000B  
Photometric Analyzer  
Part II: Analysis Unit  
NEC or ATEX Type  
5000B - GP, Rack, Panel (Integral or Remote)  
5000 - GP, Bulkhead (Z-Purged in Div II or Zone 2 areas)  
I, II, B, C, D (Integral)  
5000B-(X-Purged, 1,1, B, C, D) (Integral)  
5000B-(Purged, Zone 1) (Integral)  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 5000B Photometric Analyzer  
Table of Contents  
4 OperationalTheory  
4.0 Operations ..................................................................... 4-1  
4.1 Control Function ............................................................ 4-2  
4.1.1 Analysis Section .................................................... 4-2  
4.1.2 Explosion-Proof Version: Control Module .............. 4-2  
4.2 Start Up.......................................................................... 4-3  
4.2.1 Preliminary Inspection ........................................... 4-3  
4.2.2 Pre-Start Up Electrical Checkout ........................... 4-4  
4.2.3 Power On Observations ......................................... 4-4  
4.3 Calibration ..................................................................... 4-6  
4.4 Operational Theory Continued ....................................... 4-17  
4.4.1 Source Module....................................................... 4-17  
4.4.2 Sample Cell ........................................................... 4-19  
4.4.3 Power..................................................................... 4-22  
4.4.4 Detector Compartment........................................... 4-22  
5 Routine Maintenance  
5.0 Routine Maintenance..................................................... 5-1  
5.1 Automatic Operation and Routine Operational Duties ... 5-1  
5.2 System Visual Check and Response Procedure ........... 5-1  
5.3 Routine Maintenance..................................................... 5-2  
5.4 Suggested Preventive Maintenance Schedule .............. 5-2  
5.5 Service Procedures and Adjustments ............................ 5-3  
5.5.1 Electronics ............................................................. 5-3  
5.5.2 Power Supply Test Points....................................... 5-3  
5.5.3 Setup of the Signal Processing Front-End Amplifier.. 5-3  
5.5.4 Oscilloscope Display of the I to E Converter Output .. 5-4  
5.5.5 Balancing the Optics for Equal Light Transmission  
with Zero Fluid in the Sample Cell......................... 5-5  
5.5.6 Setup of the Logarithmic Amplifier ......................... 5-6  
5.5.7 Inverting Amplifier .................................................. 5-6  
5.5.8 Integrated Reference and Measuring Signals........ 5-7  
5.5.9 Battery-Powered Oscilloscope Synchronization Point 5-7  
5.6 Interface Board Terminal Strip ........................................ 5-7  
ii: Part II  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Part II: Analysis Unit  
Appendix  
A-1 Specifications ................................................................ A-1  
A-2 Recommended 2-Year Spare Parts List ......................... A-4  
A-3 Drawing List ................................................................... A-5  
Part II: iii  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 5000B Photometric Analyzer  
iv: Part II  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operations/Analysis Unit 4  
4.0 Operations  
Before shipment, TAI calibrates the analyzer for your application when  
feasible. Calibration data is listed in the Appendix. Prior to calibration, TAI  
checks the electronics of the analyzer and makes all of the necessary internal  
printed circuit board adjustments. Calibration is performed to determine the  
proper or close proximity zero and span settings, and also to verify that the  
analyzer response is linear.  
After calibration, TAI makes a lengthy stability check to insure that the  
analyzer performs within all specifications. We advise that you recalibrate  
your equipment as part of start-up for the following reasons:  
1. During shipment, it is possible that components have been jarred  
out of position or damaged.  
2. Your process may be of a proprietary nature. Beyond checking  
electrical stability, TAI is unable to make a meaningful calibra-  
tion of the system, requiring these adjustments to be made by  
you.  
3. Inothercasesprecisecalibrationfromtheoriginalfeasibility/  
proposal may have determined field calibration of the analyzer  
connected to your process stream is required and accuracy will  
only be obtainable upon representatiive matching and use of the  
actual process fluid background components at the operating  
physical conditions of flow, pressure and temperature of the  
sample.  
In any case, it is important that you calibrate the analyzer when it is first  
installed. Zero checks should be made routinely once or twice a week and  
span calibration should be performed weekly, at first, and monthly after  
reproducibilityisassured.  
4–1  
TeledyneAnalyticalInstruments  
A Business Unit of Teledyne Electronic Technologies  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4 Operations/Analysis Unit  
4.1  
Control Functions  
4.1.1 Analysis Section, fully explosion proof, Z or X  
purgedNemaenclosureorCenelecPurged(pending).  
The control functions for the analysis section are located on the Front  
of the Enclosure module:  
1. POWER ON/OFF: This function is hard wired into the  
enclosure for power to the control/analysis section.  
2.a  
(C-A)(1000)  
X =  
(B-A)  
For example, if the range of analysis is 20-80%, and a desirable alarm  
setpoint is 30%, then A = 20, B = 80, and C = 30  
(30-20)(1000)  
X =  
= 167  
(80-20)  
2.b RANGE:  
Control unit allows changing the range by some  
predetermined amount, such as 4X or 5X; i.e., triple range: low  
range of 0-100 ppm, mid of 250ppm and a high range of 0-500  
ppm.  
3. Mode: Programmablefunctionsforincorporatinganautomatic  
zero, span or autoranging. Autocal also possible with fluids by  
valving or by filter solenoid.. Consult control unit sections of the  
manual.  
4. Digitaldisplayt: indicates the concentration of the component  
of interest. Consult control unit section for setting up units as  
ppm,mg/l,%,etc.  
4.1.2 Explosion-ProofVersion:ControlModule  
Whenthecontrolmoduleisintegralwiththeanalysissection(singleexplo-  
sion-proofenclosure),ithasallofthesamecontrolfunctionsastheGeneral  
Purpose controlmodule,however,thebuttonsarereplacedbytwoswitches.  
Theseswitcheshavedifferentmovementstoachievethesamefunctionalityasthe  
4buttonmodel.(Consultfactoryorassurethemanualhasthecorrectversionof  
thefrontpanelcontrolswitchesandfunctions.  
4–2  
TeledyneAnalyticalInstruments  
A Business Unit of Teledyne Electronic Technologies  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operations/Analysis Unit 4  
Thecontrolunitdigitaldisplaymusthandlefunctionsforboththeanalysis  
sectionandthecontrolmodule.  
4.2  
Start-p  
Informationcontainedinthisparagraphisbasedonthepremisethatthe  
analyzerhasbeenproperlyinstalled andthatitisinoperablecondition.If  
difficultiesariseduringstart-up,itisprobablethatsomeformofdamagehas  
incurredduringshipmentorsomeinstallationerrorhasinadvertentlybeenmade.  
Accessorytestequipmentisnotnecessaryforstart-upofthe5000B..  
However, if the analyzer malfunctions at start-up, an oscilloscope and a  
multimeterwillberequiredfortroubleshooting.TAIrecommendsthatadual  
traceoscilloscopebeused.Adualtraceoscilloscopewillpermittheoperatorto  
seetwodifferentwaveformssimultaneously.  
4.2.1 PreliminaryInspection  
Beforepowerissuppliedtotheanalyzer,allmodulesshouldbeopened  
and inspected for damage or loose components. Plug-in circuit cards should  
be firmly seated in their sockets. All barrier strip wiring connections must be  
inspected, and user-installed wiring between units verified as being in agree-  
ment with the system interconnection diagram.  
ControlSettings  
Prior to turning on the power, the controls on and within the control  
module should be positioned as follows:  
Analyzers not equipped with Auto Zero  
1. ON/OFF functionality hardwired at customers remote terminal  
block or power switch.  
2. SPAN control preset to the setting noted in Specific Application  
Data in the Appendix.  
Analyzers equipped with Auto Zero  
1. ON/OFF functionality hardwired at customers remote terminal  
block or power switch.  
2. SAMPLE/ZERO switch on ZERO.  
3. SPAN control preset to the setting noted in Specific Application  
Data in the Appendix.  
4–3  
TeledyneAnalyticalInstruments  
A Business Unit of Teledyne Electronic Technologies  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4 Operations/Analysis Unit  
4.2.2 Pre-Start-upElectricalCheckout  
Afterthepreliminaryprocedureshavebeenaccomplished(refertoPrelimi-  
naryInspectionandControlSettings,above),theintegrityofthesysteminter-  
connectionandthepowersourcesmustbeverifiedbeforeattemptingtheanalyti-  
calstart-upprocedures.  
Theobservationsandmeasurementsdescribedinthefollowingparagraphs  
arevitaltotheoperationoftheanalyzer.Iftheanalyzerdoesnotrespondas  
described,theequipmenthasbeendamagedinshipmentorinstallation,orthe  
user-installedwiringisinerror.Ifaproblemarises,correctitbeforeproceeding  
further.  
Thefollowingproceduresshouldbeemployedinthesamesequencein  
which they appear.  
4.2.3 PowerOnObservations  
Turn both the analysis/control module at the remote wiring panel of  
customer to ON and make the following observations:  
1. Open the control/analysis section detector module and verify that  
the chopper motor is operating. Open the cover of the  
compartment and verify motion. The motor should start turning  
the instant power is established. If it does not, check the integrity  
of the main fuse on the control module door. If the fuse is blown,  
re-check the power service connections on TS2 in the control  
module, and the control module to analysis section  
interconnectionwiring(refertoElectricalInstallationinChapter  
3.0: Installation, and dwg. _________).  
2. The lamp source should light the instant power is established.  
Open the enclosure door and view the source module to verify  
illumination.  
WARNING: 1. The light intensity from the Mini IR lamp or quartz iodine lamp  
If used is intense and should not be looked at directly without  
specialprotectiveeyewear.Protectivegoggleswithshaded  
lenses (Fed. Spec. #5) should be worn if it is necessary to look  
directly at the source.  
2. Before opening any of the enclosure covers with the power on,  
make certain that the area has been classified as safe to do so.  
4–4  
TeledyneAnalyticalInstruments  
A Business Unit of Teledyne Electronic Technologies  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operations/Analysis Unit 4  
3. Iftheanalyzersectionhasbeenequippedwiththeautomaticzero  
SOL option, the zero solenoid device in the accessory sampling  
system should have energized the instant power was established.  
The device (or devices) should be energized because the mode  
switch has been preset to the ZERO position.  
4. If the test procedure was normal, the devices should have been  
seen or heard to operate as described by personnel located at the  
analysis section installation site, and no further check need be  
made at this time. If operation is not as described, refer to  
Troubleshooting in Chapter 5.0.  
5. Check analyzer operation by completing the calibration  
procedure described below in Calibration after allowing the  
instrument to warm up. The electrical circuits take from 30 to 40  
minutestostabilize.  
NOTE: It will take at least four hours for the analyzer to completely stabilize  
withrespecttotemperature.  
4–5  
TeledyneAnalyticalInstruments  
A Business Unit of Teledyne Electronic Technologies  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4 Operations/Analysis Unit  
4.3  
Calibration  
StandardizationFluids  
Twostandardizationfluidsarenecessarytocalibratetheanalyzer:  
1. ZeroFluid: Thezerostandardfluidmusthaveacompositionsimilar  
tothesample,andideally,containsnoneofthecomponentsof  
interest.Thezerofluidshouldbelaboratoryanalyzedtodetermineits  
composition.Theexactcompositionmustbeknown,asthe  
accuracyoftheanalysiscanbenobetterthantheknowledgeofthe  
standardizationfluid.  
2. SpanFluid: Thespanfluidmustberepresentativeofthesamplefluid  
incomposition(likethezerofluid),andcontainameasuredquantity  
ofthecomponentofinterest.Thecomponentofinterestcontentof  
thefluidshouldbeintheregionof75%to100%oftherangeof  
measurement.Aswiththezerofluid,theaccuracyofthesystemis  
dependentupontheyourknowledgeofthespanfluidcomposition.  
Ideally,thespansampleshouldcontainatleast75%ofwhateverthe  
instrumentissetuptoanalyze.  
Procedure  
1. Introduce zero fluid (see above) into the sample cell. Flow liquid  
samples into the cell at a flow rate of about 50 cc/minute to allow  
the sample to completely flush to a steady state concentration.  
Make certain that bubbles are not introduced or formed in the  
cell. (Some back-pressure may help avoid this.) Gas samples can  
be introduced at about 200-500 cc/minute.  
2. The digital meter should be capable of adjusting the zero level to  
an on scale reading. Consult section for zero functioning. If the  
reading cannot be adjusted then the signals will require  
rebalancing for your particular process fluid. The filter wheel  
located inside the detector module must be removed and balanced  
optically with screens. See section:___________  
b. If the measuring peak is so far out of balance with the  
reference peak that1 it can no longer be adjusted with  
potentiometerR3, theopticalfiltersrequirere-screening.  
If the analyzer has been in use for some time, it is possible  
that the sample cell windows need cleaning or that a filter  
hasdeteriorated.  
Anothercauseofpeakimbalancemightbethatthesample  
4–6  
TeledyneAnalyticalInstruments  
A Business Unit of Teledyne Electronic Technologies  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operations/Analysis Unit 4  
chemicalbackgroundhaschanged.Insomecases,TAIis  
notabletoduplicatethebackgroundyoursamplefor  
purposesofopticallybalancingtheanalyzerbeforeshipment.  
Inthesecases, itisnecessaryforyoutoscreenforan  
approximatebalance,andtothenelectronicallyadjustR3for  
precisebalance.Todoso:  
1. Reset R3 to its midpoint.  
2. Re-screen the filter wheel, as necessary, to obtain a  
measuring voltage within 10% of the reference voltage.  
Every time a screen is added or removed from a filter, the  
analyzer must be turned off, and the filter wheel removed  
from the analyzer. When the filter wheel is replaced in the  
analyzer, its rotational position is not critical, but the  
white backing must face in the direction of the position  
sensor. The filter wheel must also be securely tightened,  
so that no slippage or vibration can occur. See Figure 5-1.  
NOTE: Refer to Figure 5-1 when installing filter screens. Screens should be  
installed in the filter disk cup, under the filter holder. Also, if a bal-  
ance ring is used, it should be placed over the screen.  
c. After screening to bring the measuring voltage to within  
10% of the reference voltage, adjust R3 as in step a above  
to make the measuring peak voltage read the same as the  
reference peak voltage.  
5. Turn the NORM/ZERO switch to ZERO. The analysis section  
meter should be made to read zero by adjusting the coarse ZERO  
control on the analysis section power module.  
6. Check the control module SPAN setting to make sure it agrees  
with the calibration results obtained by TAI (see Specific  
Application Data in the Appendix).  
7. Adjust the ZERO control on the control module to give a zero  
reading on the control module meter.  
8. Add span fluid (see Span Fluid, above) to the sample cell. Make  
certain that all of the zero fluid is displaced.  
9. Adjust the control module SPAN control to obtain the correct  
meter reading with reference to the concentration of span fluid.  
For example, if the sample contains 100% of whatever the  
4–7  
TeledyneAnalyticalInstruments  
A Business Unit of Teledyne Electronic Technologies  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4 Operations/Analysis Unit  
instrumentissetuptoanalyze,thentheSPANcontrolmustbe  
adjustedforafullscalereading,i.e.,200ppmwaterinEDC.  
10. Re-checktheZEROsettingwithzerofluid.  
11. Ifdesired,thelinearityoftheanalyzercanbecheckedwithafluid  
intermediateinconcentrationbetweenthezeroandthespanfluid.  
12. Theanalyzerisnowcalibrated.Itisoftendesirabletocheck  
calibration (fine tune) on a dynamic sample from your process,  
double checking the 5020 analyzer response with laboratory  
analyzed grab samples. This is desirable where there is a  
possibility that your sample stream may have some background  
materials not in the makeup of the calibration fluid.  
13. From time to time, re-check the zero setting. If it is found that  
there is no zero drift, re-checking the zero setting may become  
unnecessary or may be performed only on an occasional basis.  
14. Some optical filters used in some applications will be so  
temperature sensitive that screening must be performed with the  
filtersneartheoperatingtemperature.  
DIRECTIONS FOR USING TELEDYNE CALIBRATION OPTIONS  
FOR ON-LINE PROCESS PHOTOMETRIC ANALYZERS  
I
General  
These calibration options consists of procedures for calibrating analyzers  
on-line. These techniques are offered when process conditions (high  
pressures, temperatures, sample phase changes) are extremely difficult to  
deal with and calibration standards are impossible, difficult to obtain, are  
too expensive to use due to volume consumption into the system, are  
unstable, or are difficult to introduce into the analyzer or sample system  
because of toxicity, pressure, etc. Also many times the sample stream  
composition is difficult to match with a calibration fluid.  
In one case an optical absorbing filter(s) is inserted into the light path  
of the analyzer. The absorber is actuated by a switch(s) located on  
the readout panel of the control unit. The option functions by pre-  
senting a calibrated and reproducible absorbance or fluid into the light  
path of the analyzer.  
4–8  
TeledyneAnalyticalInstruments  
A Business Unit of Teledyne Electronic Technologies  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operations/Analysis Unit 4  
A zero may be established with a reproducible simulated (offset)  
zero absorbance fluid in the sample cell of the analyzer. When  
possible this fluid should be stable for H2O content and should  
approximate the process being measured to minimize the offset  
zeroing, otherwise, dry air or nitrogen can be used. The analyzer  
must first be calibrated for sensitivity full scale by using known  
certified/analyzed samples containing the component (analyte(s) to be  
measured.  
NOTES:  
These samples (low/mid/high about 1 liter each) must be  
prepared and provided for introduction into the zero and/or  
span port of the sample system, with outlet collected from the  
cal return port of the sample system. A 500 to l liter glass  
syringe with Luor-loc to 1/4" tube connections should be used  
for introducing the samples into the analyzer sample cell for  
zero and span cal adjustments. These are introduced through  
the calibration ports usually labelled zero or span fluid inlets.  
The exiting samples when nontoxic are usually collected in  
waste containers from the cal-return port of the instrument  
cabinet. These cannot be used again as water vapor from  
containers and atmosphere will contaminate them. They would  
require analysis again before reintroducing for correlation  
purposes. Also, its is a good practice to flush copiously the  
sampling train with each calibration sample prior to accepting  
any read value. Each sample should be allowed to stabilize  
before reading. When a N2 flush is used to dry the sampling  
train before each sample is read, the readings will come to  
equilibrium faster since no water cross-contamination should  
be evident or occur.  
11  
IF CALIBRATION PROCESS REPRESENTATIVE SAMPLES  
CANNOT BE OBTAINED,  
THEN A DIRECT ON-LINE TWO-POINT CALIBRATION APPROACH  
CAN BE USED AS FOLLOWS:  
Assuming customer has designed in a sampling valve (this  
valve may contain a chromatographic septum port for a needle,  
i.e., for toxic or flammable samples at low pressure), hookup  
for a sampling bomb (high pressure), grab sample port, or  
sealed container to draw off representative process grab  
samples. Note: All samples are brought to the analytical lab  
for speedy analyses (preferably analyzed within 15 minutes) to  
correlate to process conditions. The sample analyzed should  
be by an analytical method with an acceptable precision and  
accuracy as used to normally monitor the process and confirm  
the on-line analysis. The on-line analyzer is only as accurate  
4–9  
TeledyneAnalyticalInstruments  
A Business Unit of Teledyne Electronic Technologies  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4 Operations/Analysis Unit  
as the off-line method used to verify the process variables.  
Below is an example of an on-line calibration when measuring  
0-200ppm water in liquid TDI (assumes TDI is a toxic liquid at  
standard temperature and pressure conditions).  
The first sample should be taken when the analyte (H20)  
concentration is known to be low (approximately O-.002 %  
water). The sample is quickly withdrawn and analyzed in the  
lab to determine the water content via an acceptable lab  
method (This may be Karl Fischer titration, gas chromatogra-  
phy, IR spectroscopy or volume extraction/distillation, etc.).At  
this time, the instrument zero is made to read exactly zero on  
the output of the analyzer (i.e., on the % or PPM of the panel/  
digital meter, 4-20 ma current, or 0-1 VDC etc.) using the zero  
potentiometer of the analyzer or the digital or microprocessors  
zeroing calibration adjustment, Record the lab value, ana-  
lyzed corresponding to the zero set-point adjustment.  
A second sample from the process must be grabbed after a  
steady state upscale reading has been reached. Hopefully, at  
this time, the customer can make a minor changes in the  
process to cause a slight upscale reading (1 0-30% of the full  
scale or more if possible) in order to obtain a steady-state  
reading at another concentration level. It this is not possible,  
then an appropriate time interval must occur when the process  
naturally changes to an upscale steady-state level concentra-  
tion. At this time the second calibration sample is grabbed and  
immediately analyzed. Record this value, This higher concen-  
tration value is subtracted from the first sample analyzed, Its  
difference value1 defines the slope of the calibration  
curve.  
value1- This value will be used to determine the calibration  
slope of the instrument range.  
Using the span adjustment of the analyzer, set the output reading to agree  
with this difference in concentration. Once the slope is set, adjust the zero  
of the instrument to agree with the true concentration to that of the analyzed  
steady-state sample presently flowing through the instrument cell. Now the  
analyzer is calibrated to the process fluid.  
Zero calibration using secondary standard zero and span flag(s)  
AFTER the analyzer has been calibrated on the process fluid, a simulation  
of this process on-line calibration can be made using internal synthetic  
4–10  
TeledyneAnalyticalInstruments  
A Business Unit of Teledyne Electronic Technologies  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operations/Analysis Unit 4  
optical filters. These optical filters are stable and are chosen to indicate the  
measured component at the required instrument wavelengths. Therefore,  
future, faster, yet much simpler calibration zero and span checks can be  
made without the unnecessary tedious on-line calibration described above  
nor the consumption of expensive calibration fluids which may be difficult to  
handle or obtain, are unstable, or toxic in nature.  
A Fitter(s) at the factory is chosen and installed in the instrument (approxi-  
mating in absorbance value to the anticipated stream composition back-  
ground) to give reproducible zero/ span value when nitrogen or air back-  
ground is in the measuring cell. This capability is possible when the zero  
offset between the process fluid and dry air or nitrogen remains close to the  
zero point of the actual process fluid (i.e., after calibration, the zero off-set  
can be adjusted over the entire range of the meter, plus or minus 100 %  
from zero). In the case of the calibration zero check value, the requirement  
for installing a zero optical flag with dry nitrogen as the background fluid  
depends upon how unbalanced the reference and measure channel peak  
heights are with respect to each other as measured using an oscilloscope at  
Test Point 1 (TP1 -Violet)) of the AGC- board designated B-1 4521-4. A  
good ratio (ref/meas.) estimate to keep within is (1 volt minimum: 2 volt  
maximum) to (2 volt maximum: 1 volt minimum) between the zero process  
fluid signal levels and the nitrogen calibration fluid background signal levels.  
With the process fluid in the cell at the zero point of the calibration range,  
the reference and measure pulses are optically balanced using density  
screens placed over the reference and/or measure filters. These signals are  
balanced to 9 volts each as measured at TP4 (ref-orange) and TP3 (meas-  
yellow) on the PEAK LEVEL DETECTOR BOARD designated B- 14074-  
A. The spectral background absorbance at the reference versus measure  
wavelengths of the process fluid determines the magnitude of the unbal-  
anced ref/meas ratio. This ratio is kept within the 1:2 to 2:1 values for  
proper operation of the AGC circuitry and therefore proper stability and  
signal/noise levels of the instrument output. After balancing on a zero  
sample of the process fluid, if the ratio is not found to be within the 1:2 to 2:1  
range, a zero flag is necessary and installed in the sample cell compartment  
along with a span flag if ordered. This zero flag when chosen is also  
solenoid actuated to bring the ratio within the 1:2 to 2:1 range between  
process zero fluid and zero nitrogen or dry air background. The span flag is  
actuated while the zero offset level (usually when nitrogen is used as a zero  
calibration fluid) is switched in.  
Zero calibration check after process on-line calibration.  
To check zero on the analyzer, immediately back-flush out the process fluid  
and dry the cell out with a reproducible zero fluid that should never contain  
any of the analyte. (This is usually done with N2 assuming the process fluid  
has low vapor pressure and can be purged dry in an acceptable short time  
4–11  
TeledyneAnalyticalInstruments  
A Business Unit of Teledyne Electronic Technologies  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4 Operations/Analysis Unit  
period). In this case, nitrogen or dry air (typically-100OF dewpoint or 2ppm  
water) can be used. Manually switch in the zero offset check from the  
control unit to operate the zero solenoid flag (depending upon the application,  
this may or may not simultaneously insert a zero flag into the light path;  
indeed, 90% of the time it is not needed). The factory zero offset value  
was set from its best determined factory calibration and was approximately  
set between 10-20% of the full scale range for the zero. After the field on-  
line calibration, a zero offset value around a 10-20% meter range should be  
set, using the potentiometer adjustment located within the analyzer control  
unit to bring it into the 10-20% range. This may require a short term  
area declassification (on explosion proof units) in order to make this adjust-  
ment.  
The zero offset adjustment potentiometer is panel-mount located  
within the control unit. The potentiometer identified (zero- offset) is  
adjusted while the reproducible zero fluid background is in the sample  
cell. (This zero check may be automatically or manually operated  
depending upon the sample system and electronic instrument design).  
RECORD the zero calibration reading for future reference and  
agreement checks of the zero stability.  
Note the instrument zero reading should typically be between 10-20%  
of full scale calibrated range (chosen at the factory), This reading  
should always repeat on nitrogen with the zero offset and/or  
flag introduced. If not, readjust the instrument fine or coarse zero  
potentiometer (s) or microprocessor/computer etc., to read the  
correct concentration value known (determined after the original on-  
line process calibration).  
Span calibration check after performing a process on-line calibration.  
After zero calibration above.  
To check span on the analyzer, immediately back-flush out the  
process fluid and dry the cell with a reproducible zero fluid that does  
not contain any of the analyte. In this case, nitrogen or dry air  
(typically -75 ‘C dew point or less than 1 PPM water) should be  
used. Manually introduce the span flag from the readout control unit  
to operate the solenoid containing the appropriate span filter into the  
light path of the analyzer (this may be automatically or manually  
operated depending upon the sample system and electronic instru-  
ment design). RECORD the span calibration reading for future  
calibration reference of the span sensitivity agreement. The filter  
span flag is chosen at the factory to represent an upscale reading of  
typically between 60-90% of the full scale range.  
NOTE: The zero or span flag readings should always repeat if the  
analyzer remains in calibration. Any drift in the zero or span can  
4–12  
TeledyneAnalyticalInstruments  
A Business Unit of Teledyne Electronic Technologies  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operations/Analysis Unit 4  
always be corrected using the calibration zero and/or span features.  
SUMMARY  
The zero and/or span calibration of 5000 series photometric Teledyne analyzers can be  
quickly verified manually by using Zero and/or calibration zero offsets and/or span flags.  
Refer to typical Teledyne Analytical 1475 sample system below, 3-way valves 1,2,+ 3  
below are used to check internal calibration using zero and/or span filters with N2.  
CALIBRATION OF H2O IN SOLVENTS  
EXAMPLE  
Based upon our experience when suppling photometric NIR analyzers for 0-100ppm water in solvents  
like EDC, TDI, benzene and the like, the following can be incorporated for the customers pertinent  
solvent.  
1
Customer is responsible for introducing a reproducible dry zero fluid, typically (-75deg C)  
dewpoint Nitrogen or reproducible dry EDC or TDI etc., within +/- 1 ppm H2O (for accuracy of +/- 1  
% full scale).  
2
When customer supplies a reproducible dry TDI zero fluid: A span flag option is available to  
verify instrument full scale sensitivity. This optical flag may be (manually or automatically) introduced  
into the light path within the temperature controlled (within +/- 1 deg C) sample cell compartment. This  
optical span flag ( when in this temperature stable environment simulates a very stable upscale ppm  
H2O reading on top of the DRY reproducible TDI zero background.  
3
Conversely, a zero flag and/or offset is (manually or automatically) switched into the light path  
of the analyzer after drying out the sample cell of process fluid. Consult your manual for the factor  
calibration values obtained.  
Actual concentration  
5000/5020 series analyzer meter  
Zero check  
5 ppm water/TDI  
or  
5ppm water in the 5020 meter or output. Check.  
0 ppm water in N2  
5ppm water on the 5020 meter or output. with zero flag  
offset introduction.  
delta 0ppm water or stability check on zero verified.  
Span Check  
mole-sieve or dried sample  
(could be from process)  
5ppm water in TDI  
85ppm water in TDI  
(grab sample from process)  
5ppm water on the 5020 meter or output. Zero verified.  
85 ppm water as determined by Karl Fischer titration.  
(lab technique must be verified for precision.  
(sampling device, Karl Fischer titrator and operator  
technique)  
4–13  
TeledyneAnalyticalInstruments  
A Business Unit of Teledyne Electronic Technologies  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4 Operations/Analysis Unit  
or  
85 ppm simulated water using  
span flag on top of dry liquid TDI  
delta 80 ppm agreement on slope  
sensitivity.  
or  
0 ppm water in N2  
5 ppm water on the 5000B meter  
or output with zero-flag/offset  
intro.  
0 ppm water in N2  
delta80ppmsimulatedsensitivity  
change of water.  
Note: the span flag checks the sensitivity change of the 5000B  
and not the zero drift of the analyzer. The mole sieve dried TDI  
checks the zero of the instrument provided it can be reproduce the  
zero by consistently removing all the water to the same ppm level,  
(somewhere in the 2-10 ppm water range). The long term  
efficiency of removing the water by mole-sieve contact should be  
known for the customers process fluid, otherwise the customer  
should have an acceptable off line lab technique to analyze  
prepared samples be fore intro into the 5020 field analyzer. With  
Teledyne’s approach, using N2 as the dry background for both  
zero and span checks, eliminates the concern of the mole-sieve  
being saturated with time. Indeed, the N2 Nitrogen background  
approach can be used to verify the inefficiency of the mole-sieve  
(saturation). Teledyne prefers this approach using easily  
reproducible N2 since it is much simpler to flush out the sample  
cell and check the instrument provided the original calibration on  
TDI is used to set up the zero and span instruments of the optical  
filters. See attached piping on how the appropriate valves are  
operated to switch in N2 in the reverse mode to provide a stable  
zero and or span check.  
4–14  
TeledyneAnalyticalInstruments  
A Business Unit of Teledyne Electronic Technologies  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operations/Analysis Unit 4  
1
SampleReturn  
Sample  
Flowmeter  
.2-2GPH  
Nitrogen in - dry purgeout  
of cell when manually  
zeroing/spanning  
CalibrationReturn  
Nitrogen in  
By-Pass  
Analyzer  
enclosurepurge  
0-200ccm  
Flowmeter  
2-30GPH  
PressureRelief  
Valve  
V-621  
Sample Temperature  
Compensation  
Maximum inlet pressure  
< 175 psig (12 Bar A)  
2
3
5000B  
Analyzer  
Cell  
GO  
Sample in  
Fast loop and  
By-Pass Filter  
(5 microns)  
zeroand/orspanopticalflags  
formanualorauto-calibration  
Span in  
Zero in  
ITEM  
PART NO.  
DESCRIPTION  
THIS DRAWING IS THE PROPERTY OF TELEDYNE ANALYTICAL INSTRUMENTS AND CONTAINS CONFIDENTIAL  
INFORMATION. IT IS NOT TO BE COPIED WITHOUT WRITTEN PERMISSION.  
TOLERANCE UNLESS  
OTHERWISE SPECIFIED  
SCALE  
FRAC.  
DEG.  
ANGULAR  
TELEDYNE Analytical Instru-  
ments  
DATE  
1/18/2001  
abusinessunit ofTeledyneElectronicTechnologies  
INDUSTRY, CALIFORNIA 91749  
CUSTOMER  
MODEL  
5020 SERIESNIR  
LIQUID SAMPLE SYSTEM  
Measurement(s)  
0-% or ppm H2O in solvents  
APP  
DESIGN: BY  
R. DOWNIE  
ENGINEER  
K. KOTHARI  
316SS,VITONORKALREZ,TEFLONMATERIALS  
RFQ No.  
R01  
SALES ORDER NO.  
4–15  
TeledyneAnalyticalInstruments  
A Business Unit of Teledyne Electronic Technologies  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4 Operations/Analysis Unit  
ATTACHMENT5000B/5020NIR ANALYZER  
Quote "Exceptions" and "LIQUID PHASE Conditions" for this application.  
1
Response Time is proportional to sample system design for take-off distance, process pressure, line  
size, sequencing and bypass, by-pass analyzer flow design, dead-volumes/tee's, sample cell volume  
and instrument electronics.  
2
3
Maximum cell operation pressure to be 175 psig, based upon standard Sapphire window cell design.  
Higherpressuresavailable.  
The sample flow through the analyzer should be regulated to within ± 5% of its recommended absolute  
nominal set point flow value (typically 20-60ccm). This will avoid any pressure drops across any  
orifices or poor temperature control of the sample.  
4
5
The temperature of the sample from the take-off thorough the sample cell must be non-freezing, non-  
crystallizing, non-coating as provided by customer.  
The sample temperature should not vary at the inlet more than ± 15 OC of the calibration temperature,  
norexceed the design operating specifications typically5-45 OC. Assume sample below 45 C and  
requiresnocooling.  
6
7
Any background components later found to be present and not specified in the original application  
feasibility may void the instrument performance should they interfere. particulates assumed none.  
If compounds are found present, such as moisture, acid gases or corrosive solvents, etc., the sample  
system materials may be compromised for corrosion resistance integrity. We assume that the sample  
will not polymerize, clog or react in any 1/4" sample line, filter or sample cell when transporting through  
the sample systems for all inlet ambient or sample pressures and temperatures from takeoff or derived  
from sample conditioning. Special materials may be required for stability, corrosion protection, coalesc-  
ing, filtering, hazardous protection or due to heating or cooling.  
8
9
We recommend when ambient temperatures exceed outside a 1-50 OC range that the instrument and  
sample system be conditioned appropriately for heating/ cooling without condensation, crystallizing,  
coating, polymerization, etc.  
Should the general purpose or explosion proof instrument be supplied on a back-panel or as a bulk-  
mount housing, the customer must install to protect the instrument to meet harsh ambient conditions  
for dust, water, snow, wind, corrosion, etc.  
10 Teledyne is not responsible for applying a general purpose instrument in a hazardous area or where a  
flammable or toxic gas or liquid is brought to an analyzer above its lower explosive limit and the area  
has been classified as general purpose and safe.  
11 The internal calibration features using optical filters are optimized and used when the sample can be  
flushed (backflushed usually) out of the sample cell using dry nitrogen or air to a reproducible back-  
ground zero. The zero offset and/or span flag is introduced and set-up to simulate the on-line calibration  
only after the customer has calibrated the analyzer on the representative process fluid. Thereafter, no  
calibration fluids are needed to check analyzer.  
4–16  
TeledyneAnalyticalInstruments  
A Business Unit of Teledyne Electronic Technologies  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operations/Analysis Unit 4  
4.4 Operational Theory Cont'd  
The energy source for the analyzer is most commonly provided by as a  
high intensity quartz iodine lamp located in the source module. Quartz iodine  
was chosen because it produces sufficient NIR to operate the system and  
maintains a nearly constant brightness over its lifetime. (See Figures 2-1 and  
2.2).  
This energy is then fed through the sample, which is temperature  
controlled, and into the detector module where it passes through a rotating  
filter wheel before reaching the lead sulfide (PbS) detector (PbSe also used  
for longer wavelengths (cooled and non-cooled).  
The filter wheel, driven at 30 RPS or 1800 RPM by a synchronous AC  
motor, contains two optical filters with bandpasses selected for each applica-  
tion, thus providing reference and measuring pulses from which the required  
information may be obtained.  
The detector receives pulses at the rate of 60 PPS, or two pulses per  
revolution of the filter wheel. Every other pulse is from the measuring filter,  
while the alternate pulse is from the reference filter, so that pulses through  
the measuring filter alternate with pulses through the reference filter. A filter  
position sensor, which is an optical device having an integral light source  
and light detector, differentiates between the two.  
The two entrained pulses received by the detector each revolution are  
amplified through a preamplifier which is physically located inside the sealed  
compartment with the filter wheel and detector. This signal is then sent to a  
clamping circuit where an exact zero reference is established.  
This clamped video signal is then fed through a gain control network,  
which is controlled by the automatic gain control loop, through another  
amplifier, to the electronic switch. This switch is controlled by the switch  
driver network which derives its information from the filter position sensor in  
order to separate the entrained video signal into its component parts of a  
measuring peak and a reference peak. These peaks are then fed through a  
balancing network and channeled into separate peak height detectors which  
produce DC voltage levels which are exactly equal to the peak height or  
absolute magnitude of the voltage from the base to the peak of each of the  
pulses.  
At this point the reference signal is fed back to the automatic gain  
control loop to maintain the desired system gain. In addition, both the mea-  
suring and reference levels are fed to selector switches in order to enable  
4–17  
TeledyneAnalyticalInstruments  
A Business Unit of Teledyne Electronic Technologies  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4 Operations/Analysis Unit  
direct meter indication, which greatly eases the task of balancing the system  
duringinitialsysteminstallationandperiodsofcalibration.  
TheDCvoltagelevelsarefedtoalogarithmicratioamplifierwhichpro-  
ducesavoltageoutputthatisproportionaltothelogarithmoftheratioofthetwo  
DCinputvoltages.Thisoutputvoltage,directlyproportionaltotheconcentration  
ofsample,is,withincertainlimits,alinearfunctionoftheconcentration.For  
purposesoftransmission,thevoltagesignalisconvertedbyanE-to-Iconverter;  
thus,theoutputsignalfromtheanalysisunitisacurrentsignalthatisproportional  
totheconcentrationofsampleinthesamplemodule.  
Thereisalsoanoptionofprovidinganautomaticzerofunction(seedraw-  
ingB-14729)inthecontrolunit.Thiscircuitprovideselectricalsignalsfor  
switchingafluidwhichcontainsnoneofthematerialtobemeasuredintothe  
samplemodule,electricallyadjustingthezerosandswitchingbacktosample.  
4.4.1 SourceModule  
Thesourcemoduleisthesourceofinfraredenergy.Thisisprovided  
through the use of a high-intensity mini IR lamp or a quartz iodine lamp  
operating directly from a 5.0 volt stable supply or a 6.3 V transformer. To  
ensure a stable source of radiation in the face of line-voltage variations, the  
lamp transformer derives its input directly from a line voltage-regulating  
transformer, selected for its ability to maintain a constant output voltage level  
regardless of fluctuations in the input line voltage within the control range of  
105 to 130 VAC.  
In some applications where we have an abundance of energy due to  
low sample absorption, the focusing lens is removed to avoid excess energy  
reaching the detector. However, other systems have high energy losses in the  
sample module due to strong sample absorbance or exceptionally long  
sample path-lengths. These systems require a focusing lens to gather and  
collimate the radiation for maximum utilization of source energy. The  
collimating lens is usually quartz, but may be CaF2 for longer wavelengths.  
Also the source lamp may be a differenct type, to emit more radiation at the  
longer measuring wavelengths. (Consult manual addendum descriptions).  
4–18  
TeledyneAnalyticalInstruments  
A Business Unit of Teledyne Electronic Technologies  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operations/Analysis Unit 4  
Figure 4-1. OpticalSystem  
4.4.2 Sample cell  
Thesamplecell,generallyconstructedof316SS,islocatedinthepathof  
theNIRradiation,betweenthesourceandthedetectormodules.Eachcom-  
poundinthesamplepathexhibitsitsowncharacteristicabsorptionspectrum.  
Cellspacerthicknesseswillvarydependingupontheabsorbanceofthecompo-  
nentofinterestatthemeasuringwavelength.Duetothepossiblevariationof  
absorptionwithtemperature,itiseithernecessarytomaintainthesampleata  
constanttemperatureduringanalysisortocompensatefortheminorfrequency  
shiftsofthewaterbondingwhichcanchangeduetosampletemperaturevaria-  
tions..Toachievethis,twoseparatemethodsoftemperaturecontrolareem-  
ployedforconstanttemperaturecontrol.  
1. Apreheater(heatexchangercoil)isusedontheincomingsample  
stream to raise or lower it to the desired temperature controlled  
level.  
4–19  
TeledyneAnalyticalInstruments  
A Business Unit of Teledyne Electronic Technologies  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4 Operations/Analysis Unit  
2. The entire sample module is separately controlled to maintain the  
sampletemperatureduringanalysis.  
3.Thesampleistemperaturecompensatedforbysoftwarewhileinputingthe  
temperturevalueatthesamplecell.  
Figure4-2  
AnalyzerSystem-BlockDiagram  
4–20  
TeledyneAnalyticalInstruments  
A Business Unit of Teledyne Electronic Technologies  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operations/Analysis Unit 4  
Coarse  
Ze ro  
15 VDC  
To Log Amplifier  
Control  
MEAS  
REF  
ZERO  
Se le c t o r  
Swi t c h  
To Meter Driver  
Po w e r  
5A  
115 VAC to  
So u rc e M o d u le  
115 VAC to  
Line Voltage  
Re g u l a t o r  
Transformer  
S1  
115 VAC  
Detector Module  
60 Hz  
Input  
Preheater  
Temp.  
Control  
Thermistor  
and Heater  
Sa m p le  
Module  
Temp.  
Thermistor  
and Heater  
Control  
Detector  
Module  
Temp.  
Thermistor  
and Heater  
Control  
Figure4-3  
PowerModule-BlockDiagram  
4–21  
TeledyneAnalyticalInstruments  
A Business Unit of Teledyne Electronic Technologies  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4 Operations/Analysis Unit  
4.4.3 Power  
SeeFigure 4-3.Thepowertotheanalyzerunit,enablestheswitching  
functions,andtemperaturecontrolforthedetectorcompartment.  
Withthemicroprocessorcontrolunit,thefollowinginformationisprovided.  
Figure2-3.  
PowerModule-BlockDiagram  
Thetemperatureinthedetectorcompartmentiscontrolledbyinputsfroma  
RTDdevicewhichfeedstoaPIDcontrollercapableofsetpointsoftemperature  
versustemperaturereadout.  
4.4.4 Detectorcompartment  
Afterenergyhaspassedthroughthesample,itarrivesatthefilterwheel  
whereitisfedalternatelythroughtwofilters(measuringandreference)before  
reachingthedetector.  
Thesefiltersarespeciallyselectedforeachapplicationaccordingtothe  
absorption characteristics of the compounds under analysis. The reference  
and measuring filter waveforms occur along a baseline at approximately 16  
milliseconds intervals; each reference or measuring waveform reoccurs at a  
time interval of 33 mS, or one per revolution of the filter wheel.  
At the detector, infrared energy is transformed into electrical pulses and  
fed through an impedance-matched preamplifier (see dwg. A-14619). De-  
pending upon the application, length of the cell spacer, etc., the gain of the  
preamplifier may vary from 1 to 10, depending upon the energy intensity at  
the detector, to achieve an AC signal output of approximately 0.1 to 1.0 volt  
peak-to-peak.  
Additionally,thedetector,filters,andpreamplifierarehousedinanelectri-  
callyandthermallyisolatedcompartmenttoprovidemaximumstabilityand  
minimumnoise.Thisbox,orcompartment,isnormallytemperaturecontrolledat  
46 °C.  
Asnotedpreviously,thefilterwheelisdrivenbyasynchronousACchop-  
permotorwhichoperatesat1800RPM.Thefilterwheelperformstwofunc-  
tions:(1)switchingfilters,and(2)choppingtheopticalsignaltogivepulses  
whichcanbeamplifiedforhighqualityprocessing.  
4–22  
TeledyneAnalyticalInstruments  
A Business Unit of Teledyne Electronic Technologies  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Part II: Maintenance/Analysis Unit  
Maintenance 5  
Maintenance  
5.0 Routine Maintenance  
5.1  
Automatic operation and routine operational duties  
Thesystemoperatescontinuouslywithoutadjustment. Undernormalcondi-  
tions,afteryouprogramthesystemforautomaticoperation,onlyroutinemaintenance  
procedures are necessary. The most common failure condition is a temporary  
interruptionofthepowerservingtheinstrument. Ifthepowerserviceisinterrupted,the  
sourcelampintheanalyzerwillrestartautomaticallyaslongasthereisnodefectinthe  
lampcircuitoritsstarter.  
Youcandetectalampoffconditionwiththesignalfailurealarmcircuit,butyou  
mustconnecttherelaycontactsfromthealarmtoyourindicatingdevice. Inaddition,  
youwillexperienceanalarmconditionwhenthecellwindowsareextremelydirtyorthe  
electronicsfailinthedetector-converter,logamplifier,orinvertercircuits. Whenthe  
alarmcircuitispoweredindependentlyfromtheanalyzerpowersource,thealarm  
circuitisfail-safeandwilldetectpowerfailure.  
Amessagesuchas"CellFailcheckthedetectorsignal"mightbedisplayed  
ifa1ampoffconditionoccurs  
5.2  
System Visual Check and Response Procedure  
1.  
2.  
Verifythatthesignalfailurealarmisnotinfailurecondition.  
Verifythatthezeroandspancontrolsettinghavenotbeen  
disturbed.  
Part II: 5-1  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5 Maintenance  
Model 5000B Photometric Analyzer  
3.  
4.  
Verifythatthe chartrecordercontainsanormaldisplay.  
Verifythattherecorderhasasufficientsupplyofchartpaperand  
ink.  
5.3  
Routine Maintenance  
Keep the sample lines and components, including the measuring cell  
within the analyzer sample module, free of deposits and leaks. You must  
determine the interval between cleaning procedures empirically, because the  
duration of time that the system runs without attention is related directly to the  
sample’scondition.  
5.4  
Suggested Preventive Maintenance  
Schedule  
DAILY  
1. Visuallyinspectthecompletesystemforobviousdefects,such  
asleakingtubesorconnectors.  
2.  
3.  
4.  
Verify that the sample pump (if applicable) is running.  
Verify that the signal failure alarm is not in failure condition.  
Verify that zero and span settings are correct.  
WEEKLY  
1.  
Examinesamplecellwindowsforaccumulationofsolids.  
Remove and clean as necessary.  
Calibratethesystem.  
2.  
ANNUALLY  
1.  
Checktheelectronicscalibration.  
2.  
ChecktheUVsource.  
NOTE: Be sure to wear UV filtering eye goggles.  
3.  
Checkthesolenoidvalves.  
Teledyne Analytical Instruments  
5-2: Part II  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Part II: Maintenance/Analysis Unit  
Maintenance 5  
5.5  
Service Procedures and Adjustments  
5.5.1 Electronics  
TAI aligns the system’s electronics. However, you may  
need to touch up the circuitry, using the following procedure.  
EquipmentRequired:  
Oscilloscope (dual trace is preferred, but not required) To observe  
oscilloscope test points switch the vertical input selector of the scope to DC.  
Switch to AC to observe the demodulator switch signals.  
DVM(DigitalVoltmeter)  
PC Board Extender  
Use the PC board extender whenever you need to adjust trimpot.  
Because all PC board connectors are keyed to avoid wrong positioning in the  
connectors, you must remove the key and after testing you need to replace the  
key with long-nosed pliers. Turn off the power during this operation. Never  
disconnectorconnectthePCboardswiththepoweron,becauseyoumaydamage  
the PC board C-MOS devices.  
5.5.2 Power Supply Test Points  
Measure+15volt±1voltDCand-15volt±1voltDConthedifferentialpower  
supplyPCboardinthecontrolunit. Refertothepowersupplyschematicintheback  
ofthemanualtoidentifythepowersupplytestpoints.  
5.5.3 Setup of the Signal Processing Front-End  
Amplifiers  
Fillthesamplecellwithairorastablefluid,suchthatthephotoenergy  
thatstrikesthedetectorisconstant. Astablefluidisdistilledortapwater. This  
stepmaybeomittedwhenthesystemisstableinitspresentstate.  
Part II: 5-3  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5 Maintenance  
Model 5000B Photometric Analyzer  
If you open the detector module, keep stray light out by covering the  
opening with a dense black cloth. If you do not take this precaution, the result  
is a misinterpretation of the scope patterns. On general-purpose systems, the  
scope test points are in the bottom of the detector module and are accessible  
withoutopeningthemodule.  
5.5.4 Oscilloscope Display of the I to E Converter Output  
The output of the I to E Converter is observed at the output of the second  
amplifier. Theobjectiveofthisoperationistosetuptheopticalsystemandthe  
gain of the second amplifier in such a way that the analyzer keeps operating  
within its dynamic range.  
ConnecttheoscilloscopetoTP3. Theoscilloscopedisplaysthemeasur-  
ing and reference pulses in an alternating pattern. The display is created by the  
light passing through the reference and measuring filters as they are brought in  
and out of the light beam by the rotating filter wheel. These light pulses are  
convertedtoelectronicenergywhichisamplifiedandbroughttoTP2. Thebase  
line represents the blocking of the light beam by the opaque part of the filter  
wheel.  
To identify which of the pulses is the measuring peak, insert the span  
filter(whenpresent)orapieceofflatglassorclearplasticinthelightbeam. The  
peakthatbecomestheshortest(retractsexcessively)isthemeasuringfilterpulse.  
Incaseyoucannotsetthegainproperly,becausethepeaksaretooshort,  
tootall,ortoomuchoutofbalance,adjustR2trimpotontheconverterPCboard  
until you obtain the desired peak height as observed on the scope (usually 8 to  
9 volt) for the tallest of the two peaks. Never leave the system operating with  
peaks exceeding 10 volts or you may saturate the logarithmic amplifier. You  
should not permit oscillations or distortions in the peaks.  
5.5.5 Balancing the Optics for Equal Light Transmission  
with Zero Fluid in the SAMPLE CELL  
The objective of this procedure is to obtain measuring and reference  
peakheightsasdisplayedontheoscilloscopethatareapproximatelyequal,with  
the tallest peaks set at 8 to 9 volts. This must be done with air or zero fluid in  
thecell.  
Teledyne Analytical Instruments  
5-4: Part II  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Part II: Maintenance/Analysis Unit  
Maintenance 5  
Theprocedureispurelymechanicalandconsistsofadjustingtheamountof  
lightpassingthrougheitherthemeasuringorreferencefilter,neverboth. Screens  
(wiremesh)ofvaryingdensityareusedforthisoperationandarepartofthesmall  
tookkitaccompanyingtheinstrument.  
1.  
Observe the oscilloscope and judge if optical balancing is  
needed. When the difference is less than 1 volt, balancing is not required. The  
tallest of the two peaks should be adjusted to 8 or 9 volts with the gain control  
R2onthedetectorPCboard. Whenthiscannotbedonebecausebothpeaksare  
tooshortortoolong,searchforscreensmountedinthelight path,usuallylocated  
inaholderonthelightpipewhichinterconnectsthedetectorandsamplemodule,  
and remove or add screens, as necessary.  
2.  
Whenbalancingisneeded, identifythepeaksasoutlinedunder  
Section  
3.  
For example, if the reference peak is the shorter one, stop the  
filter wheel with your hand and see if screens are located behind the reference  
filter. The reference filter is identified by the letter “R” engraved on the filter  
wheel.  
4.  
Ifscreensarefound,removethemaftertakingthefilterwheeloff  
the shaft with the special Allen wrench supplied in the tool kit.  
5.  
Afterremovalofthescreensandremountingthefilter,mountthe  
filter wheel back on the shaft. Position it correctly on the shaft by lining up the  
two paint marks on shaft and wheel.  
6.  
Turn on the instrument and observe the balance on the oscillo-  
scope.  
a.  
Ifthereferencepeakisnowtootall,removethefilterwheeland  
addascreenoflesserdensitybehindthereferencefilter. Repeatthisprocedure  
until the peaks are within 1 volt of each other.  
b. Ifthemeasuringpeakisequaltoorwithin1voltofthereferencepeak,  
the system is optically balanced and ready for calibration.  
c. If the peak is still too short, repeat the procedure, but thistime put a  
screen behind the measuring filter to shorten its peak.  
Part II: 5-5  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5 Maintenance  
Model 5000B Photometric Analyzer  
7. Afterthepeaksarebalanced,adjustthegaincontroluntilthetallestof  
thetwopeaksis8to9volts. Thepeaksshouldstillbewithin1voltofeachother.  
8. Itisalwaysgoodpracticetooperatetheanalyzerwithaslowagainas  
possible. Therefore,withthegaincontroljustbarelyoffitsstop, onceagainremove  
or add screens in the light path to obtain as high a voltage as possible without  
exceeding9voltsforthehighestpeak. Read-justgainfor8to9volts.  
Thisconcludesthebalancingprocedureandtheinstrumentisreadyfor  
calibration.  
5.5.6 Setup of the Logarithmic Amplifier  
Theamplifierisinvertingandcontinuouslytakingthelogarithmoftheoutput  
signalofthesecondamplifier. Youcanobservetheoutputbyconnectingthescope  
probe to TP4.  
The correct wave shape has a rounded negative going pulse that is the  
signalandaflat-toppedpositivepulsethatdepictssaturationofthelogamplifier.  
Youshouldnotpermitdistortionsoroscillationsintheroundedpeaks.  
Whenthepositivegoingpulseisnotflatorisdistorted,adjusttrimpotR3  
onlyenoughtoobtainaflatpositivepulse. Ifyouoveradjust,youmaylosepart  
of the second decade of absorption and affect the accuracy of analysis for high  
concentrations of the component of interest where the measuring pulse can  
become very short. The log amplifier saturates because the amplifier is  
incapableoftakingthelogarithmoftheslightlynegativebaseline.  
5.5.7 InvertingAmplifier  
Theamplifierisinvertingandhasagainof1. Itinvertstheoutputsignalofthe  
logarithmicamplifierandactsasabufferbetweenthelogarithmicamplifierandthe  
reedswitchandintegrators. Toobservethe outputoftheinverter,connectthescope  
probetoTP5. ThewavemustbeaduplicateofthatobservedonTP4,exceptthat  
itisinverted.  
Teledyne Analytical Instruments  
5-6: Part II  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Part II: Maintenance/Analysis Unit  
Maintenance 5  
5.5.8 Integrated Reference and Measuring Signals  
You can observe the reference and measuring signal at the first stage of  
integration by connecting the scope probe to TP6 (reference signal) and TP7  
(measuring signal) at the detector unit. A dual trace scope is advantageous but  
not required for this observation.  
Thetestpointssignificanceisthattheyrevealproperswitchaction. The  
displayshowsasawtoothpatternthatisacharge-dischargeofthefirstcapacitor  
intheintegratingnetwork. ThisrippleistheACcomponentofthereferenceand  
measuring signal after the pulses are converted to DC. The sawtooth patterns  
must be displayed 180° with respect to each other as viewed with a dual trace  
scope. They must both be present.  
If one is missing, the switch is not switching. If the sawtooth shows a  
broken pattern, the switching action isfeeble or irregular. Usually, you can fix  
the faulty condition of the switch by slightly changing the switch position.  
The action of a bar magnet and a rotating chopper disc activate  
themagneticmercuryreedswitch. Analuminummotormountingblockhouses  
a bar magnet. This bar magnet is parallel with the mercury chopper switch.  
The chopper disc is a green and black disc mounted on the filter wheel  
shaft next to the motor. The disc is composed of both magnetic and non-  
magneticmaterials. Astheshaftrotates,themagneticportionofthediscshorts  
themagneticfluxasitpassesbetweenthemagnetandtheswitch. Thenon-magnetic  
portionofthediscenablesfluxlinesfromthebarmagnettoactivatethemercury  
switch.  
5.5.9 Battery-Powered Oscilloscope  
Synchronization Point  
Because the line frequency cannot synchronize battery-powered oscillo-  
scopes, use TP8 at the detector module to provide external synchronization.  
5.6  
Interface Board Terminals Strip  
AtthebottomoftheinterfacePCBontheControlUnit,arethreeterminal  
strip wherewiringisdistributedtoothersectionsoftheModel6000BSystem.  
Such as AC power for the D2 lamp power supply, DC Power to the preampli-  
fier,HighDCvoltageforthephotodetector,andsignalstocontrolcalibration  
solenoidsandfilters. Togainaccesstothisterminals,thesilkscreencovermust  
beremoved. Theseterminalsarewiredinthefactory.  
Part II: 5-7  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5 Maintenance  
Model 5000B Photometric Analyzer  
WARNING: DANGEROUS HIGH VOLTAGES ARE PRESENT AT  
THESE TERMINALS. TRAINED PERSONNEL MUST  
REMOVE THE SILKSCREEN COVER ONLY. EXER-  
CISE EXTREME CAUTION.  
The firststripterminalhasthreecontactslabeledN,GandH. Thelabels  
standforNeutral,Ground,andHot. ThisistheACpowerstripterminal. It  
feedsACpowertoothercomponentsoftheModel6000BSystem, suchasthe  
D2lamppowersupply,heater,andtemperaturecontrollerPCB.  
ThesecondstripterminalhasfourcontactslabeledSHLD,SIG,GND,  
MEAS and REF. This strip terminals are dedicated to the signals coming  
fromthephotodetectoramplifier. Thelabelsstandfor:  
SHLD:Shield.Shieldformthepreamplifiercableconnectstothiscontact.  
SIGGND:SignalGround. Groundreferenceforboththemeasureandthe  
reference signal.  
MEAS: MeasureSignalvoltage.  
REF: ReferenceSignalvoltage.  
Thethirdterminalstriphaseightcontactslabeled-230VDC,+15VDC,-15  
VDC, COM, SPAN FLTR, SPAN SOL, ZERO FLTR, ZERO SOL. This  
strip feeds the high voltage needed on the cathode of the photodetector, DC  
power for the photodetector preamplifier, and control signals for the sole-  
noidsandfilters. Thelabelsstandfor:  
-230 VDC: This is the negative high voltage fed to the photodetector  
cathode, about -230 VDC.  
Teledyne Analytical Instruments  
5-8: Part II  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Part II: Maintenance/Analysis Unit  
Maintenance 5  
+15VDC: PowerSupplyvoltagefedtothephotodetectorpreamplifier,  
+15 VDC.  
-15VDC: PowerSupplyvoltagefedtothephotodetectorpreamplifier,  
-15 VDC.  
COM: Common reference to the +/- 15 VDC and the -230 VDC power  
supplies.  
SPANFLTR: Spanfiltersignal,ACvoltage.  
SPANSOL: Spansolenoidsignal,ACvoltage.  
ZEROFLTR: Zerofiltersignal,ACvoltage.  
ZEROSOL: Zerosolenoidsignal,ACvoltage.  
Part II: 5-9  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5 Maintenance  
Model 5000B Photometric Analyzer  
Teledyne Analytical Instruments  
5-10: Part II  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Photometric Analyzer, 5000B  
Appendix  
Appendix  
A-1 Specifications  
5000B Digital Control Module:  
Ranges: Four Programmable Ranges, field selectable  
within limits (application dependent) and Auto  
Ranging  
Display: 2 line by 20 alphanumeric VFD accompanied  
by 5 digit LED display  
SignalOutput: Two 0-1V DC (concentration and range ID)  
Two 4-20mADC isolated (concentration and  
range ID)  
RS232  
Alarm: Two fully programmable concentration alarm  
set points and corresponding Form C, 3 amp  
contacts. One system failure alarm contact to  
detect power, calibration, zero/span and  
sensorfailure.  
Mounting: Bulkhead Mount, NEMA-4 rated  
OperatingTemperature: 0-50oC  
Teledyne Analytical Instruments  
A-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix  
Models 5000B  
Typical Analytical Performance Specifications:  
(will vary per application)  
Accuracy:  
Noise:  
±2% of full scale or better, application dependent  
Less than ±1%  
Diurnal:  
Less than 1% per 20oF (10oC (source/detector  
dependent)  
SampleCell:  
Stainless steel with Sapphire window  
standard. Other materials available.  
.01 to 10 inches  
Cell Length:  
Flow Rate:  
50 to 1500 cc/min  
Light Source:  
Sensitivity:  
TungstenMini-Lamp,optionalQuartzHalogen  
.015 to 3 absorbance units.  
+/- 1% of scale or better  
Reproducibility:  
FilterWavelength: 750nm to 5 microns, application dependant  
Sample Pressure:  
Response time:  
Sapphire window 250 psi  
Programmable  
Other Specifications:  
Ambient temp range: 32-122 °F (0-50 °C)  
Internal calibration: Optical span flag/zero and span calibration  
NOTE: 5000B/5020B (sample temperature sensing and output correction)  
Power Rating: 115/230 VAC, 50/60 Hz,Approx 200 watts  
Readout Device: Digital, 2 line alphanumeric vacuum fluorescent display,  
(VFD)  
Analog Output Signal: 0-1 V (standard), 4-20mA isolated  
Area Classification  
Non-hazardous:Model5000B  
Hazardous / Explosion-proof:  
Model 5020  
5000B, 5020B Purge options:X, Y, Z purge, Cenelec IIB  
ANDH2(PENDING)FORZONEIORCLASSI,DIVI,B,C,D  
AREAS  
Teledyne Analytical Instruments  
A-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Photometric Analyzer, 5000B  
Appendix  
Enclosure Purging:  
Purge air startup*:40 – 80 PSI (3 – 6 BAR), 0.5 CFM (15 L / min)  
Purge optical path*:15 – 30 PSI (1 – 2 BAR), 10 cc/min, purity of  
Nitrogen, - 110 °F (-80 °C) dewpoint  
* If ordered with purge option  
Other Features:  
* Three user selectable ranges  
* Signal and Range ID output: 0-1VDC and 4-20maDC (isolated)  
* Programmable auto ranging  
* Range ID contacts, 3A@250VAC resistive  
* Two adjustable concentration alarm set points with programmable relay  
function Form C contacts, 3A@250VAC resistive  
* Programmable auto calibration with mode ID Form A normally open contacts  
* Remotely operated calibration (customer supplied valves) and 24 VDC  
signal), application dependent  
* Self diagnostics with Form C failure contacts  
* Full duplex RS232 communication link  
* Alphanumeric VF display for set up and diagnostics  
General Installation Note: Protect the instrument from  
direct sunlight and rain.  
Teledyne Analytical Instruments  
A-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix  
Models 5000B  
A-2 Recommended 2-Year Spare Parts List  
QtyP/NDescription  
C-75825A  
Motherboard, Control Unit  
Amplifier, ControlUnit  
5000B Interface PCB  
Differential Power Supply  
Measuring PCB  
C-67990  
D-67990  
1
1
1
1
1
1
A-9306  
C-40265A  
L323  
UV LAMP, 5000B  
B77609  
B74423  
C-14449  
Parabolic Reflector Assy.  
Detector assy (C78194 assy)  
TemperatureController, SampleCell  
2
2
O79  
O22  
O-Ring, Teflon  
O-Ring, Teflon  
1
5
5
1
2
1
2
F-1233  
F-57  
Span Flag Solenoid, new version  
Fuse, 5A Slo-Blo  
F-14  
Fuse, 10A Slo-Blo  
D3  
PbS detector standard  
Sample Cell Window (sapphire)  
Accessory Kit  
C128  
A-16776  
F1268  
Fuse, 6.3 A Fast-blo  
1
1
1
F1092  
F850  
F1216  
Filter, 1.9 WB  
Filter, 1.94 NB (H2O in Acetic Acid, etc.)  
Filter, 1.90 NB (H2O in EDC etc.)  
_____________________  
Note: Orders for replacement parts should include the part number (if  
available) and the model and serial number of the instrument for  
which the parts are intended.  
Orders should be sent to:  
TELEDYNE Analytical Instruments  
Teledyne Analytical Instruments  
A-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Photometric Analyzer, 5000B  
Appendix  
16830 Chestnut Street  
City of Industry, CA 91749-1580  
Phone (626) 934-1500, Fax (626) 961-2538  
TWX (910) 584-1887 TDYANYL COID  
Web: www.teledyne-ai.com  
or your local representative.  
A-3 Drawing List, 5000B  
SEE MANUAL ADDENDUM OF Q FOR COMPLETE DRAWING  
LIST.  
Teledyne Analytical Instruments  
A-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix  
Models 5000B  
Teledyne Analytical Instruments  
A-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Thermo Products Furnace PHCFA072DV4R User Manual
Tivoli Audio Cassette Player Sound System User Manual
TOA Electronics Security Camera C CV854D 3 CE User Manual
Toshiba Flat Panel Television 26AV502RY User Manual
Toshiba Network Hardware 061215 User Manual
Toshiba Personal Computer PSA20 User Manual
TP Link Network Card TL PA4010P User Manual
UNICOM Electric Log Splitter 8023af User Manual
Visioneer Digital Camera MX 230 User Manual
Weber Gas Grill 56213 User Manual