Teledyne Oxygen Equipment 9060Z User Manual

TELEDYNE ANALYTICAL  
INSTRUMENTS  
Oxygen Transmitter  
Model 9060Z  
Technical Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
5.3 KEYPAD........................................................................................................................ 28  
5.3.1 Keypad in RUN mode .......................................................................................... 29  
5.3.2 Keypad in SETUP / COMMISSIONING / CALIBRATION modes............................... 29  
6. SETUP MODE................................................................................................................. 31  
6.1 FUNCTION SUMMARY TABLE ....................................................................................... 31  
6.2 CHANGING OPTIONS ................................................................................................... 31  
6.3 SETUP MODE FUNCTIONS............................................................................................. 32  
6.3.1 Probe 1 Offset...................................................................................................... 32  
6.3.2 Lower Line Items ................................................................................................. 32  
6.3.3 Oxygen Display Units........................................................................................... 33  
6.3.4 Damping Factor ................................................................................................... 33  
6.3.5 Process Alarms .................................................................................................... 33  
7. COMMISSIONING MODE................................................................................................ 34  
7.1 FUNCTION SUMMARY TABLE........................................................................................ 34  
7.2 COMMISSIONING MODE FUNCTIONS .......................................................................... 36  
7.2.1 Internal Date / Time ............................................................................................ 36  
7.2.2 Service Date ........................................................................................................ 36  
7.2.3 Number of Probes ............................................................................................... 36  
7.2.4 Probe 1 & 2 Type................................................................................................. 36  
7.2.5 Probe 1 & 2 Thermocouple Type.......................................................................... 36  
7.2.6 Auxiliary Thermocouple Type .............................................................................. 36  
7.2.7 Transmitter Output Channel 1 & 2 ....................................................................... 36  
7.2.8 Flue Pressure Units and Value ............................................................................. 38  
7.2.9 Temperature Units............................................................................................... 38  
7.2.10 Calibration Freezes Outputs .............................................................................. 38  
7.2.11 Solenoid 1 & 2 Operation................................................................................... 38  
7.2.12 Solenoid 1 & 2 Automatic / Manual ................................................................... 38  
7.2.13 Solenoid 1 & 2 Start Time................................................................................... 38  
7.2.14 Solenoid 1 & 2 Period ........................................................................................ 38  
7.2.15 Solenoid 1 & 2 Duration .................................................................................... 38  
7.2.16 Solenoid 1 & 2 Post Freeze................................................................................. 39  
7.2.17 Oxygen Content Calibration Gas 1 & 2 ............................................................... 39  
7.2.18 Maximum Calibration Gas 1 & 2 Positive / Negative Error ................................. 39  
7.2.19 Process Alarms .................................................................................................. 39  
7.2.20 Alarm Relay 1, 2 and 3 Function ........................................................................ 39  
7.2.21 Common Alarm Relay Function.......................................................................... 39  
7.2.22 Operation of the Alarm Relays when an Alarm is Accepted................................ 40  
7.2.23 Selecting the Correct Fuel .................................................................................. 40  
7.2.24 Reference Air Pump Options.............................................................................. 40  
7.2.25 Communications Port Options ........................................................................... 41  
7.2.26 Alarm Log Clearing ............................................................................................ 41  
8. CALIBRATION MODE ..................................................................................................... 42  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
8.1 FUNCTION SUMMARY TABLE ...................................................................................... 42  
8.2 CALIBRATION MODE FUNCTIONS ................................................................................ 43  
8.2.1 Reference Voltages ............................................................................................. 43  
8.2.2 Output Channel 1 and 2 Calibration .................................................................... 43  
8.2.3 Ambient Temperature Calibration....................................................................... 43  
8.2.4 Low Oxygen Calibration ...................................................................................... 43  
8.2.5 Transmitter Output Scale .................................................................................... 43  
8.2.6 Transmitter Output Limiting ................................................................................ 43  
8.2.7 Mains Voltage Detection ..................................................................................... 43  
8.2.8 Heater SSR Fault Correction................................................................................. 44  
8.2.9 SSR Fail Protection .............................................................................................. 44  
9. ALARMS ........................................................................................................................ 45  
9.1 COMMON ALARMS...................................................................................................... 45  
9.2 SELECTABLE PROCESS ALARMS .................................................................................... 47  
9.3 ALARM RELAY OPTIONS .............................................................................................. 48  
10. INSTRUMENT CALIBRATION ........................................................................................ 49  
10.1 CALIBRATION SUMMARY ..................................................................................... 49  
10.1.1 Calibration of the Inputs.................................................................................... 49  
10.1.2 Calibration of the Outputs................................................................................. 49  
10.1.3 Probe Calibration .............................................................................................. 50  
10.2 COLD START.............................................................................................................. 50  
10.2.1 Forcing a Cold Start ........................................................................................... 50  
10.2.2 Resetting the Calibration Factors....................................................................... 50  
11. GAS CALIBRATION CHECK AND PURGE......................................................................... 51  
11.1 PURGE....................................................................................................................... 51  
11.2 CALIBRATION GAS..................................................................................................... 51  
12. SOFTWARE UPGRADES............................................................................................... 52  
13. TROUBLESHOOTING ................................................................................................... 53  
13.1 FIRST APPROACH ...................................................................................................... 53  
13.2 DETAILED FAULT ANALYSIS ....................................................................................... 53  
14. MAINTENANCE........................................................................................................... 54  
14.1 TRANSMITTER MAINTENANCE .................................................................................. 54  
14.2 CLEANING.................................................................................................................. 54  
14.3 REPLACEMENT PARTS................................................................................................ 54  
15. INDEX ......................................................................................................................... 55  
APPENDIX 1, CONSTITUENT VALUES FOR VARIOUS F.LS ................................................... 57  
APPENDIX 2, PROBE EMF TABLES ..................................................................................... 58  
APPENDIX 3, % OXYGEN SCALE TO LOGARITHMIC ............................................................ 60  
APPENDIX 4, MODBUS™.................................................................................................... 61  
4
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
Copyright TELEDYNE ANALYTICAL INSTRUMENTS — 2012  
This manual describes the transmitter firmware version 1.24, September 2012  
Neither the whole nor any part of the information contained in, or the product described in, this  
manual may be adapted or reproduced in any material form except with the prior written approval of  
Teledyne Analytical Instruments.  
The product described in this manual and products for use with it are subject to continuous  
developments and improvement. All information of a technical nature and particulars of the product  
and its use (including the information in this manual) are given by Teledyne Analytical Instruments in  
good faith. However, it is acknowledged that there may be errors or omissions in this manual. A list of  
details of any amendments or revisions to this manual can be obtained upon request from Teledyne  
Analytical Instruments. Teledyne Analytical Instruments welcome comments and suggestions relating  
to the product and this manual.  
All correspondence should be addressed to:  
Teledyne Analytical Instruments  
16830 Chestnut Street  
City of Industry, CA 91748  
Phone (626) 9341500 or (888) 7898168  
Fax (626) 9341651  
www.teledyneai.com  
Teledyne Analytical Instruments or their authorised dealers should carry out all maintenance and  
service on the product. Teledyne Analytical Instruments can accept no liability whatsoever for any loss  
or damage caused by service or maintenance by unauthorised personnel. This manual is intended only  
to assist the reader in the use of the product, and therefore Teledyne Analytical Instruments shall not  
be liable for any loss or damage whatsoever arising from the use of any information or particulars in,  
or any error or omission in, this manual, or any incorrect use of the product.  
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
Important Notice Regarding 9060 Probe Option FIL3  
WARNING: The only identifiable standard for flame arresters for general use is British Standard BS  
EN 12874:2001. British Standard BS EN 12874:2001 refers to an operating environment up to 150  
Degrees Centigrade.  
The FIL3 device optionally fitted to 9060 Heated Zirconia Probes (the “Probes" or "Probe")  
operate in an environment considerably greater than 150 Degrees Centigrade.  
Therefore, we know of no Australian, British, European or USA standard applicable to flame  
arresters or their testing above 150 degrees Centigrade. Consequently, the FIL3 device cannot be  
certified as a safety device.  
The probe is only on probes during the start up processes of a combustion appliance.  
The Teledyne Analytical Instruments Burner Interlock Relay facility, which is a standard part of the  
Teledyne Analytical Instruments transmitter, is designed to be wired to the main safety shutoff  
fuel valves in a way that can shutdown the probe heater when the fuel valves are closed.  
The risk of ignition of flammable gas mixture at the hot end of the Probe can only be minimized by  
correct use, maintenance and operation of the FIL3 device. The user of the FIL3 device is  
responsible for verification and maintenance and correct use and operation of the FIL3 device.  
THE USER AGREES THAT IT USES THE PROBE AND THE FIL3 DEVICE AT ITS SOLE RISK. TELEDYNE  
ANALYTICAL INSTRUMENTS TO THE FULL EXTENT PERMITTED BY LAW, GIVES NO WARRANTIES OR  
ASSURANCES AND EXCLUDES ALL LIABILITY (INCLUDING LIABILITY FOR NEGLIGENCE) IN RELATION  
TO THE PROBE AND THE FIL3 DEVICE.  
The user must ensure that it correctly follows all instructions in relation to the Probe and FIL3  
device, correctly understands the specifications of the Probe and FIL3 device and ensures that the  
Probe and FIL3 device are regularly inspected and maintained.  
FIL3 equipped Probes should be inspected at least once a year for corrosion and more frequently  
if there is any reason to suspect that corrosion may have occurred.  
6
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
1. IMPORTANT NOTICES  
This manual is intended as a supplement to the 9060Z Operators Manual. It is intended to be used by  
technical personal that are qualified to install, commission, service and calibrate electronic industrial  
control equipment.  
Please read the safety information below before connecting power to the transmitter.  
CAUTION 1  
The probe heater is supplied with mains voltage. This supply has electrical shock danger to  
maintenance personnel. Always isolate the transmitter before working with the probe. The EARTH  
wires (green/yellow) from a heated probe must ALWAYS be connected to earth.  
CAUTION 2  
Combustion or atmosphere control systems can be dangerous. Burners must be mechanically set up  
so that in the case of equipment failure the system cannot generate explosive atmospheres. This  
danger is normally avoided with flue gas trim systems by adjustment so that in the case of failure the  
appliance will not generate CO in excess of 400 ppm in the flue. The CO level in the flue should be  
measured with a separate CO instrument normally an infrared or fuel cell type.  
CAUTION 3  
The oxygen probe is heated to over 700°C (1300°F) and is a source of ignition. Since raw fuel leaks can  
occur during burner shutdown, the transmitter has an interlocking relay that removes power from the  
probe heater when the main fuel shutoff valve power is off. If this configuration does not suit or if it  
is possible for raw fuel to come into contact with a hot oxygen probe then the Model 9060Z  
transmitter with a heated probe will not be safe in your application. An unheated probe can be  
utilized in such applications, however the oxygen readings are valid only above 650°C (1200°F).  
CAUTION 4  
The reducing oxygen signal from the transmitter and the associated alarm relay can be used as an  
explosive warning or trip. This measurement assumes complete combustion. If incomplete  
combustion is possible then this signal will read less reducing and should not be used as an alarm or  
trip. A true excess combustibles analyzer, normally incorporating a catalyst or thermal conductivity  
bridge, would be more appropriate where incomplete combustion is possible. Also read the probe  
electrical shock caution in the probe heater interlock caution in chapter 4.7  
CAUTION 5  
FIL3 filter. If the optional FIL3 has been fitted to the 9060H probe in this installation, please read the  
Important Notice regarding probe option FIL3 on the previous page.  
CAUTION 6  
The heater is supplied from the mains power directly, and the temperature is controlled at 720°C  
(1330°F). The outside of the process end of the probe can get to temperatures that are dangerous to  
touch. Wear insulating gloves when handling a probe that has been on.  
CAUTION 7  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
Please note that if this equipment is not installed and used in the manner described in this manual then  
the safety protection provided by the equipment may be impaired.  
2. INTRODUCTION  
2.1 9060Z TRANSMITTER  
The Teledyne Analytical Instruments 9060Z oxygen transmitter is designed to measure the oxygen gas  
concentration in combustion applications such as power stations and industrial boilers. The  
measurement is used to control the air / fuel ration in the burner to maintain efficiency and safe  
operation conditions.  
The transmitter is designed to be used in conjunction with a Teledyne Analytical Instruments probe  
model 9060H or 9060UL/UH. These two probes use a zirconia sensor that must operate at >650°C.  
The 9060H probe has an integrated heater, and the 9060UL/UH probe uses the process heat to heat  
the sensor. There is a unique cable connection between the 9060Z transmitter and the  
9060H/9060UL/UH probe.  
The model 9060Z oxygen transmitter is based on the well known model 9060H transmitter. It includes  
a number of hardware and software improvements such as a graphic display, larger characters, faster  
microprocessor, simplified set up menu, alarm logging, faster probe heater control and more  
calculated values such as burner efficiency.  
The 9060Z Oxygen Transmitter has a variety of userselectable functions. They are simple to use  
because each selection is menu driven. For options you are not sure about, read the manual on that  
particular item in chapter 6, Setup Mode.  
Features include:‐  
Inputs  
Two zirconia oxygen probes, heated or unheated  
Oxygen read 01% to 0100%  
Furnace, kiln or flue thermocouple, field selectable as type K, J, R or S  
Main flame established safety interlock (for heated probes only)  
Purge pressure or flow switch  
Outputs  
Two linearised 420mA or 020mA DC isolated outputs, max. load 1000Ω  
The output function and the range are field selectable  
Common alarm relay  
Three other alarm relays with selectable functions  
Digital Interface  
RS232 or two wire RS485 MODBUS for connection to a computer/DCS/PCL for diagnostics of the  
transmitter, probe or combustion process.  
Display  
Multi font graphical display  
Large font characters for the oxygen on the top line  
Selectable lower line items for the secondary display functions. ie Probe temperature, Oxygen second  
probe Alarm display mode that shows the time of the alarm, the acceptance time and the time that  
the alarm was cleared of up to 4000 alarm events  
2.2 SERIES 9060H OXYGEN PROBES & SENSORS  
8
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
Teledyne Analytical Instruments series 9060H oxygen probes and sensors employ stateoftheart  
zirconia sensors and advanced materials, which provide the following benefits:  
Improved control due to fast response time to typically less than four seconds  
Costefficient design provides improved reliability  
Longerlife probes with greater resistance to corrosion from sulfur and zinc contaminants in  
flue gas  
Low cost allows maintenance by replacement  
Reduced probe breakage due to greater resistance to thermal shock and mechanical damage  
during installation and startup  
Series 9060H probe or sensors are simple to install and maintain. All models provide direct  
measurement of oxygen level. Online automatic calibration check is available if required. Probes or  
sensors may be used with Teledyne Analytical Instruments oxygen transmitters and some model  
transmitters from other manufacturers.  
All Teledyne Analytical Instruments oxygen probe or sensors are designed and manufactured to  
exacting standards of performance and reliability. Series 9060H probe or sensors are the result of  
extensive research and development by Teledyne Analytical Instruments, industry and government  
agencies. Teledyne Analytical Instruments provides application and after sales support for oxygen  
probes, sensors and transmitters, worldwide.  
Model 9060H Heated Oxygen Probe  
Model 9060UL/UH Unheated Oxygen Probe  
2.3 Warning Symbols  
Danger, high voltage. Risk of electrical shock.  
Caution hot surface.  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
Caution, risk of danger. See additional information in the manual.  
3.1 HARDWARE SPECIFICATIONS  
3.1.1 TRANSMITTER SPECIFICATIONS  
Number of oxygen probes: 2 maximum  
01% to 0100%  
Oxygen range:  
Oxygen accuracy:  
±1% of actual measured oxygen value with a repeatability of ±0.5% of  
the measured value  
Type K, J, R and S  
Thermocouple types:  
Temperature accuracy:  
Analogue outputs:  
±2°C  
020mA or 420mA field selectable  
Active outputs (Do NOT loop power these outputs)  
1000 ohm max  
Output load:  
Alarm relays:  
4
Alarm relay contacts:  
2A/240VAC, 2A/30VDC  
(WARNING: Do not use both mains voltage and low voltage connections to adjacent alarm contacts)  
Mains voltage supply:  
Overvoltage:  
Power:  
100 to 240VAC, 6 +10%, 50/60 Hz  
Category II (IEC603644443)  
5W for controller plus probe power  
530W max., 25% duty cycle each probe on 240VAC  
110W max., 100% duty cycle each probe on 110VAC  
2.5A max  
Fuses:  
3A, fast blow, 250v, 20x5mm (heater fuses, 2 of)  
1A, slow blow, 250v, microfuse (PCB mtg fuse, 1 of)  
Operating Temperature 25°C to 55°C  
Relative Humidity 5% to 95% (noncondensing)  
2000m maximum  
Environmental Rating:  
Altitude  
Degree of Protection:  
IP65  
IP54 with internal reference air pump  
Case Size:  
Weight:  
260mm (10.2”) wide, 160mm (6.3”) high, 90mm (3.5”) deep  
3 Kg (6.6 lbs.)  
WARNING: All signal level connections onto the transmitter must be treated as safety extralow  
voltage (SELV) as defined in the standard IEC61140. Double insulation must be used when  
connecting these terminals to systems that might carry high voltage.  
3.1.2 STANDARD ‘U’ LENGTHS  
9060H  
9060UL/UH  
500 mm (20”)  
750 mm (30”)  
1000 mm (40”)  
1500 mm (60”)  
250 mm (10”)  
350 mm (14”)  
500 mm (20”)  
750 mm (30”)  
1000 mm (40”)  
1500 mm (60”)  
2000 mm (80”)  
Ordering Information  
10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
Probe insertion length (from process end of mounting thread to probe sensing tip.  
3.1.3 SERIES 9060H PROBES SPECIFICATIONS  
MODEL  
Application  
9060H  
Combustion flue  
9060UL/UH  
Combustion flue gases  
above 700°C (1290°F) with  
no contaminants.  
eg. natural gas, light oils  
700 to 1400°C  
gases below  
900°C (1650°F)  
Refer to note 1  
0 to 900°C. Refer to note 2  
(32 to 1650°F)  
Temperature Range  
Length  
(1470 to 2550°F)  
250 to 2000 mm  
(10” to 80”)  
500 to 1500 mm  
(20” to 60”)  
Process Connection  
1 ½ BSP " or NPT  
¾ BSP " or NPT  
Electrical Connection Weatherproof plugin connector or optional screw terminals. The plug  
connector can be supplied with the cable.  
Cable  
Order a specific length with the transmitter  
Heater  
Yes  
No  
Thermocouple  
Response Time  
Head Temperature  
K, integral  
Typically < 4 secs.  
R, integral  
Typically < 1 sec  
25 to 100°C (15 to 210°F) with weatherproof connector  
25 to 150°C (15 to 300°F) with screw terminals  
Ambient air, 50 to 150 cc/min (3 to 9 scim). Pump can be supplied with  
transmitter.  
Reference Gas  
Ref Air Connection  
Filter  
1/4 NPT "  
Integral air line in probe cable.  
Barbed fitting to 3/16" ID PVC tube.  
Removable sintered titanium alloy particulate filter, 30 micron standard,  
optional 15 micron available. Refer to note 2  
Calibration Check Gas 1/8 NPT female "  
1/8 NPT female "  
Connection  
Weight  
2 kg (4.4 lb) plus  
165 g (5.8 oz) / 100 mm  
(4”) length  
1 kg (2.2 lb) plus  
100 g (3.5 oz) / 100 mm  
(4”) length  
Notes: 1. Care must be taken to avoid contact with explosive or inflammable gases with 9060H  
heated probes when hot. Teledyne Analytical Instruments transmitters have built in safety  
protection.  
2. Process gas temperature must be below 800°C if the filters are fitted.  
Please contact factory for corrosives other than sulfur or zinc. We can provide test materials to try  
in your atmosphere.  
11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
3.1.4 OXYGEN PROBE MODEL SELECTION GUIDE  
Heated probestemperature range 0900°C (1650°F).  
*Note: (1) The Inconel option has all inconel wetted parts except for the ceramic sensor and viton  
‘o’ rings.  
12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
Unheated probes for clean gasestemperature range 7001400°C (12902550°F).  
*Note: (2) A standard oxygen probe for carburising furnaces, has a 253 MA sheath.  
13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
3.2 OPERATIONAL SPECIFICATIONS  
Range of outputs:  
Function  
Minimum  
Maximum  
Linear oxygen #1  
Very low oxygen #1  
Linear oxygen #2  
Very low oxygen #2  
Average oxygen  
0 to 1%  
0 to 0.001%  
0 to 1%  
0 to 0.001%  
0 to 1%  
0 to 100% (**)  
0 to 2.000% (**)  
0 to 100% (**)  
0 to 2.000% (**)  
0 to 100% (**)  
Logarithmic oxygen  
Reducing oxygen  
Oxygen deficiency  
Aux temperature  
Combustibles  
0.1 to 20%  
0 to 1x1030%  
10 to 0%  
0 to 100°C  
0 to 0.5%  
Fixed range  
0 to 100% (**)  
10 to 20% (**)  
0 to 1600°C  
0 to 2.0% (**)  
Carbon dioxide  
0 to 5%  
0 to 100% (**)  
Probe #1 sensor EMF  
Probe #2 sensor EMF  
Burner efficiency  
No output  
0 to 100mV  
0 to 100mV  
0 to 20.0%  
Disables the output  
0 to 1500mV (Fixed zero)  
0 to 1500mV (Fixed zero)  
0 to 100.0% (**)  
** (Adjustable zero)  
Range of local indication:  
01% to 0100%  
0.01 to 10,000ppm, automatically defaults to % range above  
10,000ppm (1%)  
Local display, lower line secondary display items:  
Function  
Range  
Probe #1  
Probe #2  
Probe #1  
Probe #2  
Probe #1  
Probe #2  
temperature Up to 1760°C (3200°F)  
temperature Up to 1760°C (3200°F)  
EMF 30 to 1350mV  
EMF 30 to 1350mV  
sensor impedance 0 to 300k Ω  
sensor impedance 0 to 300k Ω  
01% to 0100%  
Oxygen probe #2  
Average oxygen, probe #1 and probe #2  
Aux temperature  
Ambient temperature  
Ambient RH  
01% to 0100%  
Up to 1760°C (3200°F)  
20° to 70°C (5° to 125°F)  
5 to 95%  
Carbon dioxide  
Combustibles  
0 to 2.0%  
0 to 2.0%  
Oxygen deficiency  
Burner efficiency  
10.0 to 20.0%  
0 to 100.0%  
14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
The oxygen deficiency output can be used in the same way as a combustibles analyzer to signal the  
extent of reducing conditions of combustion processes.  
4. INSTALLATION AND COMMISSIONING  
4.1 MOUNTING THE TRANSMITTER  
Surface mount the transmitter case on to a flat surface or bracket, using the four mounting brackets  
provided. The transmitter should never be mounted so that it is directly expose to the sun or rain.  
Always leave at least 10cm of clearance around the four sides of the case.  
Make sure the temperature of the case is below 55°C, and that the radiated heat from furnaces and  
boilers is kept to a minimum. There should be adequate ventilation to maintain ambient temperature.  
Install the cables through cable glands. There are 4 holes cut in the base of the transmitter case; 2x  
17mm & 2x 21mm. Use a sharp knife to cut away the covering film for only the glands that are  
needed.  
NOTE: ALWAYS LEAVE THE UNUSED GLAND HOLES SEALED.  
Ensure that the electrical connection complies with the local electrical requirements. (see chapter 4.6,  
Electrical Connections)  
4.2 INSTALLING A 9060H OXYGEN PROBE  
Weld a BSP or NPT socket to the flue in a suitable position for flue gas sensing. For the correct size of  
socket refer to probe data in chapter 3.1.3. The closer to the source of combustion the smaller will be  
sensing lag time, allowing better control.  
The probe has a typical response time of less than four seconds, so most of the delay time is normally  
the transit time of the gas from the point of combustion to the point of sensing.  
15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
Probes can be mounted at any angle. However, if the probe is to be mounted on a vertical duct wall, it  
is better to angle the probe (approx. 15°) down to avoid process condensation inside the cold end of  
the probe. If there are any particulates in the flue gas, a filter can be omitted by pointing the probe  
vertically downwards. Otherwise the transmitter can be configured to automatically purge the filters  
or they can be replaced periodically.  
,
Oxygen Probe Mounting  
CAUTION  
It is important that there is no air in leakage upstream of the oxygen sensing point otherwise  
there will be a high oxygen reading.  
If the probe is to be installed on a bend in the flue, it is best located on the outer circumference  
of the bend to avoid dead pockets of flue gas flow. While the standard 9060 probe with a ‘U’  
length of 250 mm (10”) will suit most low temperature flue applications, it is occasionally  
necessary to have a longer probe with the sensing tip in the centre of the flue gas stream.  
Although it is rare, occasionally a probe may sense oxygen vastly differently from the average  
reading in the flue gas. If it occurs, then the probe should be moved, or a longer probe  
installed. This phenomenon is normally caused by stratification of the flue gas.  
16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
4.3 INSTALLING THE AUXILIARY THERMOCOUPLE  
Weld a 1/2 inch BSP mounting socket to the flue within about 300 mm (12”), and upstream of the  
oxygen probe. The thermocouple should be of similar length to the oxygen probe to prevent flue  
temperature distribution errors.  
The thermocouple should be connected to terminals 7 & 8. These terminals will not be available for an  
auxiliary thermocouple if a second probe has been installed.  
4.4A SHIELD CONNECTIONS  
All external wiring to the 9060Z transmitter should be shielded. Do not connect shields at the field  
end. Simply clip off and insulate. There are two M4 earth screw terminals available in the 9060Z  
transmitter. An extra terminal strip may be required to connect all shields together. This should be  
supplied by the installer.  
4.4B EARTH CONNECTION (PE)  
The PE (protective earth) input connection must be made to the earth stud on the right hand side of  
cabinet. The PE input connection must be the first connection onto the earth stud and it must be  
secured by a separate nut and spring washer. All other earth connections (bonding connections) can  
be made on either of the two earth studs in the base of the cabinet.  
The transmitter MUST be securely earthed.  
Assembly of the PE and bonding connections on the earth stud  
4.5 ELECTRICAL CONNECTIONS  
All wiring should comply with local electrical codes. The printed circuit boards are fully floating above  
earth. The incoming safety earth (PE) must be connected to the primary earth stud in the right hand  
side of thebase cabinet.  
The local fuse for the mains power supply, the isolation switch and the supply wiring must all comply  
with the electrical safety codes and must only be installed by qualified technicians. All earth and shield  
connections should be connected to the earth screws inside the case.  
17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
Connection Diagram for 9060Z Transmitter and one or two 9060H Heated Sensors  
18  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
Connection Diagram for 9060Z Transmitter and one or two 9060UL/UH Unheated Probes  
19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
• The mains power must be either 100/110VAC or 220/240VAC with a mains frequency of 50 or 60Hz  
• The supply circuit must be fused to at least 10Amps and have a dual pole isolation switch within  
easy access of the oxygen transmitter. The isolation switch must be marked as the isolation switch  
for this equipment. It is recommended that a separate isolation switch be used for each transmitter  
so that a transmitter can be serviced individually.  
• The power supply cables must be supplied and installed according to local regulations  
• The earth connection must comply with the local regulations must have a current carrying  
capability  
equal or greater than the supply fuse current rating  
• The earth connection must be connected to the primary earth stud inside the transmitter on the  
right hand side  
• All other bonded earth connections from the external wiring must be connected to the primary  
earth stud.  
All operations relating the electrical wiring and installation must be carried out by qualified persons in  
accordance with the safety regulations and the wiring rules.  
NOTE: The power switch in the transmitter can be used to turn off the transmitter. There must be an  
approved isolation device installed to provide complete isolation of the mains power to the  
transmitter. The mains wiring terminal must not be used as an disconnect device.  
4.6 HEATER INTERLOCK RELAYS  
CAUTION  
Explosion protection for heated probes is achieved by switching the power to the probe heater off  
whenever the main fuel valve is closed.  
The principle of safety is that if the main fuel valve is open then main flame has been established. With this  
primary source of ignition on, the probe heater can be safely switched on. The most dangerous situation is if  
fuel leaks into the combustion appliance when the fuel valve is closed. When power is removed from the  
main fuel valve the heater should also be switched off.  
To achieve this protection, connect a main fuel valve voltage free contact to the ‘BURNER INPUT’ terminals  
10 & 11. When the main fuel valve is open, the voltage free contact should be closed. For installations  
where there is no risk of explosion, connect a link between terminals number 10 & 11.  
Heater Supply Interlock Connection for Heated Probes  
If a safety interlock is not required, a wire must be connected between terminals 10 &11 to enable –  
• The heaters on heated probes  
• Process alarms  
• Autopurge and autocal checking  
• Run time timer  
4.7 CONNECTING AN OXYGEN PROBE CABLE  
Connect the probe lead as shown in the following drawings. Unheated probe leads have integral  
reference air tube. An adaptor has been supplied to connect this tube to quarter inch flexible PVC  
tubing, from the air pump or reference air supply.  
20  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
Connection of Probe Cable for Unheated Probes Models 9060UL/UH.  
Connection of Probe Cable for Heated Probes Model 9060H .  
21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
4.8 CONNECTING THE AUXILIARY THERMOCOUPLE (OPTIONAL)  
For 9060H heated probes, the auxiliary thermocouple must be a separate TC with the junction  
isolated from earth, mounted near to and upstream of the oxygen probe. It can be either a K, J, R or S  
type thermocouple. It is optional. If the auxiliary temperature or combustion efficiency is not to be  
displayed or transmitted, then an auxiliary thermocouple is not necessary. The thermocouple should  
be connected to terminals 7 & 8. These terminals will not be available for an auxiliary thermocouple if  
a second probe has been installed.  
4.9 CONNECTING THE OUTPUT CHANNELS  
The two 4 to 20 mA DC output channels are capable of driving into a 1000Ω load. The output  
terminals are 12 & 13 for channel 1, and 14 & 15 for channel 2.  
4.10 CONNECTING THE ALARMS  
A common alarm, which should be connected for all installations initiates on alarm functions  
described below. Three additional alarm relays are available for selectable functions as listed in  
chapter 9.2. Each relay has normally open contacts. The contacts will open in alarm condition. Relays  
are connected as follows:  
Relay Common Alarm  
Terminal Numbers 26 & 27  
Alarm 1  
Alarm 2  
Alarm 3  
28 & 29  
30 & 31  
32 & 33  
Common Alarms All of the following conditions will cause a common alarm ‐  
Probe 1 (2) Heater Fail  
Probe 1 (2) High Impedance  
Probe 1 (2) TC Open Circuit  
Aux TC Open Circuit  
Ref Air Pump Fail  
ADC Calibration Fail  
Alarm Log Fail  
Heater 1 (2) SSR Failure  
Heater SSR Leakage  
Probe 1 (2) Filter Blocked  
Gas 1 (2) Calibration Error  
BBRAM Fail  
Output 1 (2) Failure  
The watchdog timer is a special alarm. There will not be an alarm message displayed, but the  
transmitter will reset.  
Alarms can be accepted by pressing the alarm button (viewing the alarm messages).  
Alarm relay 2 to 4 Select any one or all of the following for each relay. Refer chapter 9.  
In addition, any of the selections that are removed from the common alarm relay can be added to  
relays two to four.  
High oxygen 1 (2)  
Low oxygen 1 (2)  
Very low oxygen 1 (2)  
Oxygen deviation  
Probe 1 (2) temperature low  
Calibration check 1 (2) in progress  
Probe 1 (2) purge in progress  
4.11 CONNECTING THE AUTOMATIC PURGE AND CALIBRATION CHECK SYSTEM  
CAUTION  
The purge and calibration solenoid valves are supplied with mains voltage. This supply has  
electrical shock danger to maintenance personnel. Always isolate the transmitter before working  
with the purge and calibration solenoid valves.  
22  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
The online auto purge and calibration check system is optional. For details on its operation see  
chapters 4.24, 4.25 and chapter 11.  
To automatically sense a blocked probe filter, a flow switch should be installed in the ‘purge’ line to  
the probe ‘CAL’ port. It should be adjusted so that it energizes just below the purge flow with new or  
clean filters installed. The flow switch contacts should be connected to terminals 16 & 17 (FLOW SW).  
If the filter is still blocked or partly blocked after an auto purge cycle, the flow switch will not energize  
and will therefore cause a ‘Probe1 (2) Filter Blocked’ alarm.  
After installation the purge/cal system should be tested thoroughly for leaks. Any leaks can cause  
significant errors if the flue is at negative pressure. If the flue is at positive pressure, an outward leak  
can cause corrosion in the purge/cal system piping and fittings.  
If probe/filter purging is required but a “Probe1 (2) Filter Blocked” alarm is not required, do not install  
the flow switch but link terminals 16 &17.  
Automatic Purge &  
Calibration check  
System  
Wiring  
Schematic  
Automatic Purge &  
Calibration check  
System  
Piping  
Schematic  
4.12 CONNECTING REFERENCE AIR  
For oxygen probes, a 1/4” tube connector on the transmitter should be connected via a nylon, copper  
or stainless steel tube to the ‘REF’ connector on the probe. If two probes are being used, a “T” fitting  
must be supplied to provide reference air supply to both probes.  
4.13 CONNECTING THE TRANSMITTER TO A MODBUS™ NETWORK  
The transmitter can be networked to other transmitters and to a network master. The network uses  
the transmitter’s RS485 port. Up to 31 transmitters can be connected to the network, and can be  
interrogated by the Network Master.  
NOTE: Hardware Protocol Selection The communications port of the 9060Z can be configured to  
communicate on either RS232 or RS485. If the transmitter is to be used on a MODBUS network, the  
9060Z transmitter must be set to RS485. For further details see chapter 7.2.25, Communication Port  
Options.  
NOTE: Terminating Resistor There is a terminating 100 ohm resistor fitted to the 17321 PCB. Link LK3  
(TERM), in the bottom lefthand corner of the PCB, is used to connect the terminating resistor. Link  
LK3 must be removed on all transmitters except the transmitter on the end of the network line. If the  
23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
network line from the transmitters is taken from the middle of the transmitter network string, a  
terminating resistor should be enabled with LK3 at each end of the network line.  
The MODBUS™ protocol of the network is –  
Baud Rate  
Parity  
19,200  
Even  
Stop Bits  
RS485  
Mode  
1
2wire (half duplex)  
RTU (binary mode)  
For more details see Appendix 5.  
4.14 ONNECTING POWER  
Before commissioning the probe, sensor or transmitter, read the CAUTION paragraphs at the front of  
this manual. The green & yellow EARTH wire MUST be connected to earth.  
4.15 COMMISSIONING RUN MODE  
When the transmitter is turned on it will go to RUN mode. The SETUP/RUN button will toggle  
between the two modes. The upper line of the display will now read ‘% OXYGEN’. If the probe or  
sensor temperature is not above 650°C (1200°F), a “Probe Low Temperature” message is flashed on  
the lower line. The probe or sensor temperature can be checked on the lower line of the display.  
4.16 PROBE OR SENSOR CALIBRATION  
The zirconia sensor provides an absolute measurement of oxygen partial pressure. There are no  
calibration adjustments, apart from ‘Probe 1 Offset’, for the probe. The zirconia sensor EMF is either  
correct or it needs to be replaced. To check that the probe is functioning correctly, first check that the  
24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
high impedance alarm, ‘Probe 1 (2) High Impedance’, is not active. The actual impedance can be  
displayed on the lower line. It should be less than 9 KΩ at 720°C (1320°F).  
Once it has been established that the impedance is normal, the offset may be set using the millivolt  
value marked on the oxygen probe. See chapter 6.3.1, Probe Offset. The probe offset can be tested on  
site. A small flow of air must be admitted to both the ‘REF’ and ‘CAL’ ports when testing the probe  
offset. If the probe is in the process, the air must fully purge the probe sensor without interference  
from the process gas sample. Teledyne Analytical Instruments probes can easily achieve this with or  
without a probe filter and a gas flow of only 1 to 5 SLPM (2 to 10 SCFH) for a 9060H probe and up to  
20 SLPM (40 SCFH) for an unheated probe. When a stable oxygen reading has been achieved, read the  
‘Probe EMF’. Enter this value into the setup menu #1 for the probe 1 (setup #2 for the second probe  
if it is installed).  
4.17 FILTER PURGING  
Purging probe filters is controlled from the ‘GAS/PURGE’ buttons on the transmitter when in ‘RUN’  
mode. If ‘Automatic’ has been selected in either ‘Sol. 1 Auto/Man’ or ‘Sol. 2 Auto/Man’ in  
Commissioning Menus 21 or 30, pressing the GAS/PURGE button will start the automatic cycle.  
Pressing the button again will cancel the auto purge cycle. If ‘Automatic’ was not enabled, the  
solenoid will only stay open for as long as the button is pressed. Gradually adjust the purge air supply  
regulator, increasing the pressure until sufficient flow is obtained to clear the filter. This is best  
checked with a dirty filter after a period of operation, by withdrawing the probe from service and  
watching any build up on the filter being blown off at the set pressure. Normally 30 kPa (5 psi) is  
adequate but the air pressure may be set as high as 100 kPa (15 psi).  
4.18 CALIBRATION GAS CHECK  
If the installation has a filter purge facility, set this up first. Refer to the previous paragraph. Press the  
‘GAS/PURGE 1 or ‘GAS/PURGE 2’ button while in ‘SET UP’ mode to obtain a reasonable flow through  
the calibration check gas flow meter. If air is being used as a calibration check gas, use the air from the  
regulator for filter purge. Then, when setting up a gas for calibration checking, set the pressure from  
the calibration gas cylinder so that it is the same as the pressure set on the air regulator. Then the  
setting on the rotameter / flow regulator will be the same as that for the airflow. The flow required is  
1 to 5 SLPM (2 to 10 SCFH) for a 9060H probe and up to 20 SLPM (40 SCFH) for an unheated probe.  
Air is not the best gas for calibration checking on a zirconia sensor. The output of a zirconia sensor  
with air on both sides of the sensor is zero millivolts. It is better to choose a gas value which provides  
a reasonable output from the sensor and which is near to the process oxygen level. A cylinder with 2%  
oxygen in nitrogen is a commonly used calibration gas. The maximum pressure on the calibration  
check gas cylinder regulators is 100 kPa (15 psi).  
Note: If ‘Dual probe’ was selected in commissioning menu 4, ‘Cal Gas 2’ must be connected to probe 2.  
4.19 DUST IN THE FLUE GAS  
For unheated probes with no filter, entrained solids or dust in the flue gas does not present a  
problem unless the dust, when settled, is not porous. Allow the dust in the process to build up on  
the probe. It will form a porous layer slowing the response time. To avoid mechanical abrasion of  
the electrode material in installations with unheated oxygen probes, pack ‘SAFFIL’ or equivalent  
alumina based ceramic fiber in the sensing holes to protect the electrode. Do not use silica based  
ceramic fibers such as ‘KAOWOOL’, which can attack the electrode at high temperatures. Once the  
dust has built up the response time of the probe will be slower.  
For heated probes the preferred method of mounting for dustladen applications is facing vertically  
25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
downwards with the filter removed. Probes can also be mounted horizontally with no filter with some  
dusts. An occasional automatic back purge is helpful in this case.  
Normally heated probes are supplied with filters for applications with particulates in the flue gas. The  
probe response time should be tested when the probe is first installed, and then regularly until it  
remains constant for a significant period. Filter purging should be set up on the time periods  
determined by these tests. To test the probe response time, use a stopwatch to obtain the time for a  
probe to achieve a 63 % change from one reading to another. If a probe filter blocks completely in a  
short period of time, then there is no option but to use the probe without the filter. A trial probe with  
filter can be installed to test whether a filter blockage is likely to occur.  
4.20 STRATIFICATION  
If the transmitter and probe have been fully tested and the oxygen readings in the flue gas are  
incorrect, gas stratification may be occurring. The phenomena cannot be anticipated for any particular  
installation. Generally, large flues have oxygen differences of approximately one percent across the  
flue. Occasionally an oxygen error of several percent may occur in a flue of any size. Moving the probe  
to a new location normally solves this problem. The effects of stratification can be reduced by using  
two probes and averaging the two oxygen readings. This can be achieved within a Teledyne Analytical  
Instruments 9060Z transmitter controlling two probes.  
4.21 CONNECTING A PRESSURE TRANSDUCER  
If the process gas pressure varies more than 4" WG and therefore requires automatic compensation,  
connect a pressure transducer as shown below. Place a link across LK2, near to the input terminals 24  
& 25. If the link is not populated, you can achieve the same thing connecting a 120Ω resistor across  
the two terminals. A pressure change of 4" WG will cause a change in the calculated oxygen reading  
of approximately 1%.  
There are no calibration adjustments for the zero or span of the pressure transducer input. Set the  
transducer scale range in the commissioning menu See Section 7.2.8. The pressure can be displayed  
on the lower line by enabling the corresponding option. See Section 6.3.2.  
26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
5. DISPLAY AND KEYPAD  
Operator interaction with the oxygen transmitter is done through a graphical display and 8 keys on  
the front of the case. The five LED indicators are used to show the status of the transmitter and to  
alert the operator of any errors.  
Each of the keys has a dual function;  
The BLACK text printed on the key is the function while the transmitter is in the RUN mode  
The WHITE text printed on the key is the function in one of the three configuring menus, (SETUP,  
COMMISSIONING and CALIBRATION)  
The menu driven modes are accessed by pressing, and in some cases holding down, the SETUP key  
(see chapters 6, 7 and 8 for details of these modes). The transmitter will return to the RUN mode  
when the SETUP key is pressed again, or if 60 seconds has elapsed since the last key was pressed  
.
5.1 Graphical Display  
In RUN mode the 9060Z shows the first oxygen probe measurement in large characters at the top of  
the display and a user selectable lower line in smaller characters below. Other items on the display  
include the activity heartbeat indicator in the top right corner, a row of single letter action indicators  
in the bottom left corner and the current time in the bottom right corner.  
In SETUP mode the RUN display is replaced with a menu driven interface. All other functions of the  
transmitter (reading, calculating and transmitting etc) aside from the keypad and display will continue  
to operate as normal.  
27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
5.2 OXYGEN DISPLAY UNITS  
The measured oxygen content displayed on the top line of the transmitter for probe 1 and on the  
lower line for probe 2 can be shown in either % or parts per million (ppm). See chapter 6.3.3 Oxygen  
Display Units for how to change the oxygen display.  
The format of the oxygen display changes to maintain the best resolution for the measurement.  
If ‘Oxygen %’ is selected the transmitter will display oxygen in the following format:  
Range  
Display format  
##.#%  
##.##%  
#.###%  
30.0 to 100.0%  
1.00 to 29.99%  
0.100 to 0.999%  
< 0.100%  
scientific notation (#.## x 10 ## % )  
If ‘Oxygen ppm’ is selected the transmitter will show the oxygen in ppm format between 10 and  
9,999ppm. Outside of this range the oxygen will be shown in % format described above.  
Display format for ‘Oxygen ppm’:  
Range  
Display format  
> 10,000ppm (1%)  
1,000 to 9,999ppm  
100.0 to 999.9ppm  
10.00 to 99.99ppm  
< 10ppm  
% format (see ‘Oxygen %’ section above)  
#,### ppm  
###.# ppm  
##.## ppm  
scientific notation (#.## x 10 ## % )  
5.3 KEYPAD  
There are eight keys built into the decal on the outside of the door of the 9060Z transmitter. The key  
functions are written in BLACK and WHITE to identify the function of the key in either RUN mode or  
SETUP mode.  
Key text  
SETUP / RUN  
RUN mode (WHITE text)  
Enter SETUP mode  
SETUP mode (BLACK text)  
Return to RUN mode  
DISPLAY / FUNCTION UP  
DISPLAY / FUNCTION  
DOWN  
Display Next Lower Line Item  
Display Prev. Lower Line Item  
Next Function  
Previous Function  
ALARM / OPTION UP  
ALARM / OPTION DOWN  
ALARM ACCEPT / ENTER  
Next Alarm  
Previous Alarm  
Acknowledge Displayed Alarm  
Next / Increment Option  
Previous / Decrement Option  
Save Current Option  
GAS 1 PURGE 1 / SENS IMP Gas 1 / Purge 1 manual activate Check Probe impedance  
Gas 2 / Purge 2 manual activate  
GAS 2 PURGE 2 / AUTO CAL  
Manually Activate Output Calibration  
5.3.1 Keypad in RUN mode  
When the transmitter is turned on and has completed its startup procedure, it will enter RUN mode  
and enable the keypad. The key functions in this mode are as follows;  
28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
SETUP / RUN key  
Pressing this key will put the transmitter into one of the menudriven SETUP, COMMISSIONING or  
CALIBRATION modes (see chapters 6, 7 and 8 for details). The function of all the keys will then change  
to the functions that they have in the SETUP mode.  
DISPLAY UP / DOWN keys  
The display keys are used to scroll the lower line up and down through the variety of measurements  
that are available on the lower line. For a complete list of options see chapter 6.3.2, Lower Line Items.  
ALARM UP key  
If there is either a new alarm or an active alarm the ALARM UP key can be pressed to examine the  
alarm status. The alarm light will be flashing if there is a new alarm or steady if there is an existing  
alarm. For more details on the alarm mode and keys see chapter 9, Alarms. The SETUP light will flash  
slowly to show that the transmitter is now in the alarm display mode.  
ALARM DOWN key  
When the transmitter is displaying active alarms (the ALARM UP key has been pressed), the ALARM  
DOWN key and the ALARM UP key allow the operator to examine the date / time of the alarm and the  
date / time that the alarm was acknowledged.  
If the transmitter was in RUN mode when this key is pressed it will go into the alarm log display mode.  
See chapter 9 for more details.  
ALARM ACCEPT key  
The ALARM ACCEPT key is used to accept a new alarm (see chapter 9, Alarms).  
GAS 1 / PURGE 1 and GAS 2 / PURGE 2 keys  
These two keys are used to turn on the gas / purge solenoids. When the transmitter is in the manual  
cal / purge mode (Commissioning functions #21 and 30) the solenoid will be activated for as long as  
the key is pressed. When the transmitter is in the auto cal / purge mode the automatic cal / purge  
cycle is started. The cycle can be stopped by pressing the same key again.  
5.3.2 Keypad in SETUP / COMMISSIONING / CALIBRATION modes  
NOTE: In order to enter either COMMISSIONING or CALIBRATION modes, the corresponding DIP  
switch on the 17321 PCB must be turned on.  
From RUN mode, if the SETUP / RUN key is pressed once, the transmitter will go into the SETUP mode.  
From RUN mode, if the SETUP / RUN key is pressed and held for 2 seconds, the transmitter will go into  
the COMMISSIONING mode.  
From RUN mode, if the SETUP / RUN key is pressed and held for 4 seconds, the transmitter will go into  
the CALIBRATION mode.  
The following key functions are then available in all of the above modes.  
SETUP / RUN key  
Pressing the SETUP / RUN key while in any of the set up modes will return the transmitter to the RUN  
mode.  
29  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
FUNCTION up and FUNCTION down keys  
These two keys allow for the selection of the setup function. A function summary table is found at the  
start of the relevant SETUP, COMMISSIONING or CALIBRATION mode chapter.  
OPTION up and OPTION down keys  
These two keys allow for changing the option for the selected function. A list/range of options for  
each function is found in the function summary table at the start of the relevant SETUP,  
COMMISSIONING or CALIBRATION mode chapter.  
ENTER key  
The ENTER key sets the currently displayed option and stores the value in nonvolatile memory to be  
retrieved on device start up. If the ENTER key is not pressed when a new option is chosen, the  
previous option will be retained.  
Probe impedance key  
When this key is pressed the transmitter will measure the impedance of oxygen sensor(s) attached to  
the transmitter. If the burner is not enabled (terminals 10 & 11) or the probe temperature is below  
700°C (1292°F) impedance checking will not be performed. A ‘Z’ will be seen in the bottom left hand  
corner of the display as the test is performed.  
Auto calibrate key  
When this key is pressed the transmitter will calibrate the analogue output channels that are set to  
auto calibration (see chapter 10). This is performed by directing the output current away from the  
output terminals (terminals 12&13 and 14&15) back to an analogue input to the transmitter in order  
to calculate a zero and span calibration factor for each of the output channels.  
During this process normal output to the analogue channels will be interrupted sending the outputs to  
zero mA.  
30  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
6. SETUP MODE  
The SETUP mode is mode is accessed by pressing the SETUP key momentarily so the words ‘Setup  
Mode’ appear at the bottom of the display. The transmitter will return to the RUN mode when the  
SETUP key is pressed again or 60 seconds after the last key press.  
6.1 FUNCTION SUMMARY TABLE  
When the transmitter is in the SETUP mode the SETUP light will be on and the words “Setup Menu”  
will be shown at the bottom of the display.  
The following table shows the SETUP functions:  
Menu #  
Function name  
Probe 1 Offset  
Probe 2 Offset  
Lower Line Items  
Oxygen Display Units  
Damping Factor  
Process Alarms  
Range  
Default value  
0.0mV  
0.0mV  
1
2
3
4
5
6
6.0 to +6.0mV  
6.0 to +6.0mV  
See Chapter 6.3.2  
Oxygen % / Oxygen ppm  
No Damping to 10 Samples  
Enabled / Disabled  
Oxygen %  
5 Samples  
Disabled  
6.2 CHANGING OPTIONS  
The purpose of having a menu driven SETUP mode is to allow for configuration of the transmitter  
using the graphical display and keypad.  
Once an option is changed and entered using the ENTER key that value immediately becomes active.  
It is also stored into the nonvolatile memory within the BBRAM and will be loaded again at device  
startup. To reset the transmitter configuration defaults see chapter 10.2, Cold Start.  
To change an option in the SETUP menu system:  
1. Select the SETUP mode by pressing the SETUP / RUN key once. The SETUP light will come on and  
the display will have the format shown below. The operations of the keys are now the operations  
written in WHITE on the keypad. The menu name is written at the bottom of the display.  
2. When the SETUP mode has been selected the required function can be found by using the  
FUNCTION UP and FUNCTION DOWN keys.  
The options available for that function can be seen by using the OPTION UP & OPTION DOWN  
keys.  
3. When the required option is on the display press the ENTER key to save that value.  
When finished, press the SETUP / RUN key to return to the RUN mode.  
31  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
6.3 SETUP MODE FUNCTIONS  
6.3.1 Probe 1 Offset  
Each Teledyne Analytical Instruments probe has an offset calibration value printed on a tag that is  
attached to the probe when it is dispatched. To achieve accurate measurement the offset value must  
be entered with the same polarity as it is printed on this label. The offset value should be within  
±1.0mV  
NOTE: An offset of 1.0mV will change the oxygen reading by approximately 1% oxygen when the  
probe is in ambient air. However, as the process oxygen measurement drops, this offset will have a  
reduced affect. At a process gas oxygen concentration of 2%, the 1.0mV offset error will only change  
the reading by 0.1% oxygen.  
If in any doubt about the correct offset value, set it to 0.0mV.  
The function ’Probe 2 Offset’ will only appear if the transmitter has been configured for 2 probes. (see  
chapter 7.2.3 Number of Probes )  
6.3.2 Lower Line Items  
This function allows the operator to change the items that are available to be displayed on the lower  
line of the transmitter in RUN mode. If the word “Enabled” appears on the display for a selected lower  
line option, the measurement will be available to be shown on the display in the RUN mode by  
scrolling through the list using the DISPLAY UP and DISPLAY DOWN keys.  
A lower line measurement can be “Enabled” or disabled by pressing the ENTER key.  
32  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
NOTE: An asterisk (*) on the end of the line identifies the item is enabled by default after a COLDSTART.  
6.3.3 Oxygen Display Units  
The top line of the display that shows the oxygen measurement can be displayed in either % or ppm  
(see chapter 5.2 Oxygen Display Units)  
6.3.4 Damping Factor  
The oxygen measurement can be damped by averaging successive readings from the probe. This will  
smooth out fluctuations in the process gas level and will slow down the reaction time of the  
transmitter. The larger the number selected here, the more successive readings are averaged and the  
smoother the measurement will be. The damped oxygen value is also used in the calculation of all  
other parameters that are based on the oxygen value.  
6.3.5 Process Alarms  
This function allows the operator to enable and disable process alarms. The process alarm menu  
functions will also be hidden if the process alarms are disabled. (See also chapter 7.2.19, Process  
Alarms)  
33  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
7. COMMISSIONING MODE  
This chapter describes the functions available when the COMMISSIONING mode is entered on the  
transmitter.  
The COMMISSIONING mode is accessed by pressing and holding the SETUP key for approximately 2  
seconds until the words “Commissioning Menu” appears at the bottom of the display. The transmitter  
will return to the RUN mode when the SETUP key is pressed again or 60 seconds after the last key  
press.  
Changing options in the COMMISSIONING mode is the same as the SETUP mode. See chapter 6.2,  
Changing Options.  
7.1 FUNCTION SUMMARY TABLE  
When the transmitter is in the COMMISSIONING mode the SETUP light will be on and the words  
“Commissioning Menu” will be shown the bottom of the display.  
The following table shows the COMMISSIONING menu functions:  
34  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
35  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
7.2 COMMISSIONING MODE FUNCTIONS  
7.2.1 Internal Date / Time  
The date and time are used in the transmitter to run time based operations such as impedance  
readings and gas calibration checks. The correct date and time should be entered by pressing and  
holding the OPTION UP & DOWN keys. The date and time will change faster the longer the OPTION  
key is held.  
7.2.2 Service Date  
The service date can be used to keep a record of when a probe was changed. The RUNTIME timer  
which keeps a track of the hours and minutes that the transmitter and probe has been operating is  
reset when the service date is changed.  
The service date and the runtime timer can be displayed as lower line items in RUN mode.  
7.2.3 Number of Probes  
The 9060Z transmitter can operate in single or dual probe mode. By selecting single probe, all dual  
probe options and alarms will be disabled and second probe options hidden from the user.  
If you are using the transmitter as a single probe device you should set the number of probes to  
‘single probe’ to disable all probe 2 alarms and display items.  
7.2.4 Probe 1 & 2 Type  
The probe type function allows a selection between a heated probe and an unheated probe types.  
Heater control and alarms will be disabled if unheated probe type is selected.  
7.2.5 Probe 1 & 2 Thermocouple Type  
The thermocouple type can be set to K, J, R or S. The thermocouple in the Teledyne Analytical  
Instruments 9060H heated probes are always K type. 9060UL/UH unheated probes are available in R  
type or S type. Other thermocouple options are made available for special installations.  
7.2.6 Auxiliary Thermocouple Type  
When the transmitter is operating as a single probe mode, the probe 2 thermocouple input (terminals  
7&8) is used as an auxiliary thermocouple input. In dual probe mode this option is unavailable.  
Use of an auxiliary thermocouple allows for basic monitoring of temperature as well as enabling the  
calculation of combustion efficiency. The thermocouple type can be set to K, J, R or S.  
7.2.7 Transmitter Output Channel 1 & 2  
The 9060Z transmitter has two fully configurable 420mA analogue outputs. The channels can be  
configured independently to output one of several calculated values.  
36  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
The zero and span of the selected output are set in the following two menus (functions 11&12 and  
14&15).  
NOTE:  
Regarding channel output configuration; both 420mA channels are separately isolated and output  
variables and scaling are individually configured for each channel.  
In single probe mode both channel 1 and 2 have identical options, however when the analyzer is  
configured as dual probe then channel 1 will output process variables relative to probe 1 and  
channel 2 will output process variables relative to probe 2.  
37  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
7.2.8 Flue Pressure Units and Value  
The 9060Z transmitter is capable of fixed or variable input pressure compensation in the calculation of  
oxygen from 1 Atm to +3 Atm. If the probe is running in a pressurized environment this value should  
be set to allow for accurate oxygen measurement. If the flue pressure is constant select 'Fixed" in this  
function and set the pressure units and value in the proceeding menus. If the pressure varies, select  
"Variable", and connect a pressure transducer to terminals 24 & 25. (See section 4.23). Set the range  
of the transducer using a zero and span value in the commissioning menus.  
7.2.9 Temperature Units  
The display on the transmitter can be changed to show temperature in either Celsius or Fahrenheit  
scales.  
7.2.10 Calibration Freezes Outputs  
During gas calibration checks it may be required that the analogue outputs remain frozen at their last  
reading. When cal freezes is enabled both analogue outputs remain frozen until the entire calibration  
cycle has completed.  
7.2.11 Solenoid 1 & 2 Operation  
The two solenoid outputs can be individually configured to perform gas calibration check or probe  
purge.  
Some of the menus between functions 22 to 28 (31 to 37) will not appear depending on what has  
been selected in function #20 and #21 (29 and 30).  
7.2.12 Solenoid 1 & 2 Automatic / Manual  
When set to automatic the solenoid will perform a preprogrammed gas calibration or purge cycle.  
When set to manual the solenoid is controlled directly from the corresponding key on the front of the  
case. When set to automatic the calibration check or the purge cycle can be started by pressing the  
corresponding key on the front of the case. It can be stopped by pressing the same key again.  
7.2.13 Solenoid 1 & 2 Start Time  
For automatic gas calibration checking or purge events a start time is specified to allow precise control  
over the timing of such events. Starting at this specified time, the next gas calibration or purge event  
will occur in successive intervals from this time.  
7.2.14 Solenoid 1 & 2 Period  
This option specifies the period between automatic gas calibration or purge events. For periods less  
than 24 hours the period is divisible into 24 hours forcing scheduling of events to occur at the same  
time each day between 1 minute and 7 days.  
The ‘No Timed Operation’ option accessible by scrolling the option below 1 minute allows for pre‐  
configured solenoid cycles to be triggered manually via the keypad, but without the automatic  
scheduling.  
7.2.15 Solenoid 1 & 2 Duration  
Duration of time that an automatic gas calibration or purge event energizes the solenoid  
38  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
7.2.16 Solenoid 1 & 2 Post Freeze  
Duration of time the transmitter waits before resuming live readings following a gas calibration or  
purge event. (See also chapter 7.2.10)  
7.2.17 Oxygen Content Calibration Gas 1 & 2  
When configured as a gas calibration check solenoid this value specifies the oxygen content of the  
calibration gas.  
7.2.18 Maximum Calibration Gas 1 & 2 Positive / Negative Error  
Specifies the maximum positive & negative error levels used during the gas calibration check.  
If the difference between the specified oxygen content and the measured process gas exceed these  
thresholds, a gas calibration error alarm will be triggered immediately.  
7.2.19 Process Alarms  
The oxygen measurements are continuously monitored by the transmitter for process alarm  
conditions (Process alarms must be enabled in the Set up menu). The limits and the alarm time delays  
are configurable for the Very Low Oxygen alarm, Low Oxygen alarm, High Oxygen alarm and the  
Oxygen Deviation alarm.  
7.2.20 Alarm Relay 1, 2 and 3 Function  
There are 3 user configurable alarm relays. Any of the 3 relays can be configured to be triggered on  
the following alarm conditions. In addition, any of the alarm conditions that are disabled from the  
common alarm relay can also be configured to trigger these 3 relays. Multiple selections can be made.  
Process Alarms  
Oxygen 1 Low  
Oxygen 2 Low **  
Oxygen 1 Very Low  
Probe 1 Temperature Low  
Probe 2 Temperature Low **  
Calibration 1 In Progress  
Oxygen 2 Very Low ** Calibration 2 In Progress **  
Oxygen 1 High  
Purge 1 In Progress  
Oxygen 2 High **  
Oxygen Deviation  
Purge 2 In Progress **  
7.2.21 Common Alarm Relay Function  
The common alarm relay can be configured to be triggered on any of the following:  
Instrument Alarms  
Probe 1 Heater Fail  
Reference Air Pump Fail  
Heater SSR Leakage  
Probe 2 Heater Fail **  
Reference Air Pump Overload Probe 1 Filter Blocked  
Probe 1 High Impedance  
Probe 2 High Impedance **  
Probe 1 Thermocouple Open Circuit  
Alarm Log Fail  
Output 1 Fail  
Output 2 Fail  
Probe 2 Filter Blocked **  
Gas 1 Calibration Error  
Gas 2 Calibration Error **  
Probe 2 Thermocouple Open Circuit ** Heater 1 SSR Fail  
Auxiliary Thermocouple Open Circuit Heater 2 SSR Fail  
** Are only available in dual probe mode.  
39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
All of the items in this list are selected as the default setting. Any of these items can be disabled from  
the common alarm by pressing the ENTER key, and they will then appear on the list in the other alarm  
relays.  
7.2.22 Operation of the Alarm Relays when an Alarm is Accepted  
An option is available to in the commissioning menu to change the operation of the alarm relay when  
an alarm has been accepted. The states for the alarm relay contacts are shown in the following table –  
Alarm state  
Accepted Relay hold settingAccepted Relay hold setting‐  
Enabled  
Disabled  
No alarm condition  
New alarm  
All alarms accepted  
Alarms self cleared  
Closed circuit  
Open circuit  
Open circuit  
Closed circuit  
Closed circuit  
Open circuit  
Closed circuit  
Closed circuit  
The default setting is ‘Enabled’ after a cold start.  
7.2.23 Selecting the Correct Fuel  
The 9060Z transmitter can calculate several measurements that are based on the oxygen reading and  
the temperature of the process gas. These calculations also require some details about the chemical  
makeup of the fuel. The additional calculations are:  
Oxygen deficiency  
The volume of oxygen required to bring the combustion back to  
stoichiometry  
Combustibles  
This theoretical scale is equivalent to a combustibles analyzer  
The calculated maximum level of CO2 in the combustion gas  
Carbon dioxide  
(maximum dry)  
Combustion  
efficiency  
Requires a stack thermocouple and so is only available in single probe  
mode  
Select the required fuel in function 50 by using the OPTION keys and then press the ENTER key. The  
table of values in the next 6 functions will be updated for that fuel.  
7.2.24 Reference Air Pump Options  
Normally the reference air is supplied from the transmitter using the internal pump. The default  
option is 5.00v.  
Function  
Options  
57  
58  
59  
Reference Air Pump  
Internal Pump Voltage  
Reference Air RH%  
External or Internal  
Pump voltage setting, between 2.5 and 5.00 in 0.25v steps  
If external is selected, set the RH level. (5% if instrument  
air is used)  
40  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
7.2.25 Communications Port Options  
The 9060Z transmitter has a serial communications port available at terminals 18 to 22. The default  
protocol is for RS232 running at 19,200 baud rate with 8 bits, Even parity and 1 stop bit.  
Function  
Options  
60 Serial Interface  
RS232 or RS485 (Use RS485 for MODBUS)  
61 Serial Baud Rate 9600, 19200, 38400, 57600 (RS232 up to 19200 only)  
62 Serial Parity  
Even, Odd or No Parity  
MODBUS  
Address  
Set the MODBUS address for this device between 1 and 246.  
0 is to disable the MODBUS  
63  
7.2.26 Alarm Log Clearing  
Every alarm that is instigated internally in the transmitter or as a process level is recorded in the alarm  
log with the activation time, accepted time and the cleared time. The last 4000 events will be  
recorded and then the oldest alarms will be dropped off as new ones occur. If you would like to delete  
the log recording select CLEAR in SETUP step 64 by using the OPTION keys and press the ENTER key.  
The message “Alarm Log Cleared” will be displayed.  
41  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
8. CALIBRATION MODE  
This chapter describes the functions available when the CALIBRATION mode is selected on the  
transmitter.  
For specific information about calibrating the transmitter see chapter 10, Instrument Calibration.  
The CALIBRATION mode is accessed by pressing and holding the SETUP key for approximately 4  
seconds until the words “Calibration Menu” appears at the bottom of the display. The transmitter will  
return to the RUN mode when the SETUP key is pressed again or 60 seconds after the last key press.  
Changing options in the CALIBRATION mode is the same as the SETUP mode. See chapter 6.2 Changing  
Options.  
8.1 FUNCTION SUMMARY TABLE  
When the transmitter is in the CALIBRATION mode the SETUP light will be on and the words “Commissioning  
Menu” will be at the bottom of the display. The following table shows the CALIBRATION menu functions:  
Menu  
#
Function name  
Range  
Default  
value  
1
2
3
4
5
Reference Voltage 1, 50mV  
Reference Voltage 2, 200mV  
Reference Voltage 3, 1200mV  
Reference Voltage 4, 2500mV  
Output Channel 1, Calibration  
40.00 to 60.00mV  
47.14mV  
180.00 to 210.00mV  
1150.0 to 1250.0mV  
2400.0 to 2550.0mV  
182.24mV  
1221.8mV  
2489.2mV  
Auto Calibrated Manual Calibrated Set Auto  
4mA Set 20mA  
3.00 to 5.00mA  
19.00 to 21.00mA  
Calibrated  
4.00mA  
20.00mA  
6
7
8
Output Channel 1, 4mA Trim  
Output Channel 1, 20mA Trim  
Output Channel 2, Calibration  
Auto Calibrated Manual Calibrated Set Auto  
4mA Set 20mA  
Calibrated  
9
Output Channel 2, 4mA Trim  
Output Channel 2, 20mA Trim  
Ambient Temperature Sensor Offset 10.0°C to 10.0°C  
Low Oxygen Cal 1  
Low Oxygen Cal 2  
3.00 to 5.00mA  
19.00 to 21.00mA  
4.00mA  
20.00mA  
0.0°C  
100.00%  
100.00%  
420mA  
10  
11  
12  
13  
14  
80.0% to 120.0%  
80.0% to 120.0%  
420mA / 020mA / 420mA over‐  
Transmitter Output Select  
scale limit / 020mA over scale limit  
15  
Transmitter Output Limiting  
Disabled / Hold 0mA / Hold 4mA /  
Hold 20mA  
Hold 0mA  
16  
17  
18  
Mains Voltage Detection Override  
Automatic / 220/240 / 10/120  
Automatic  
Automatic  
Normal  
Mains Frequency Detection Override Automatic / 50Hz / 60Hz  
Heater SSR Selection Normal  
Heater1 <> Heater2 Heaters <> Cal  
Purge  
19  
SSR Fail Protection  
Enabled / Disabled  
Enabled  
42  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
8.2 CALIBRATION MODE FUNCTIONS  
8.2.1 REFERENCE VOLTAGES  
The calibration of the analogue inputs is based on the voltage of a temperature compensated voltage  
reference integrated circuit. There are 4 voltages generated from the standard reference voltage.  
They will vary by about 1% from one transmitter to another but can be trimmed by setting the actual  
voltages into CALIBRATION menu functions #1 to 4. These voltages should be measured and the  
reference voltage entries in the menu checked every year.  
8.2.2 OUTPUT CHANNEL 1 AND 2 CALIBRATION  
The output 420mA calibration can be set either AUTOMATICALLY (default) or MANUALLY. If Auto  
Calibrated is selected the transmitter will go through an output calibration cycle when the power is  
turned on or when the AUTO CAL button is pressed in SETUP mode. This will divert the outputs back  
to the input and automatically set the 4mA and 20mA calibration.  
If Manually Calibrated is selected then the 4mA and 20mA calibration must be set in the next two  
functions. For the full explanation see chapter 10.1.2.  
8.2.3 AMBIENT TEMPERATURE CALIBRATION  
The ambient temperature measurement is used as the cold junction temperature for the  
thermocouple measurements. Use an independent temperature sensor to measure the temperature  
inside the transmitter case near the screw terminal #1. Enter this temperature into the calibration  
function 11 by using the OPTION keys and then the ENTER key.  
8.2.4 LOW OXYGEN CALIBRATION  
The low oxygen calibration factors can be used to fine tune the oxygen calculation at low oxygen  
readings. It will not affect the measurement at 20.9%. They are included to allow oxygen probes made  
by other manufacturers to be used on a Teledyne Analytical Instruments transmitter. The default is  
100% but can be set to between 80 and 120%.  
8.2.5 TRANSMITTER OUTPUT SCALE  
The transmitter outputs can be configured to either 420mA or 020mA. The two scaled outputs are  
capable of outputting 024mA and will transmit overscale unless the overscale limit option is  
selected.  
8.2.6 TRANSMITTER OUTPUT LIMITING  
If the oxygen reading is invalid the output can be made to be set to 0mA, 4mA or 20mA. An invalid  
reading is when the probe temperature is below 650°C (1200°F) or the probe thermocouple has been  
detected as open circuit. The default is to set the output to 20mA but the function can also be  
disabled.  
8.2.7 MAINS VOLTAGE DETECTION  
The default setting for the 9060Z transmitter is that it will automatically detect the mains voltage and  
frequency in order to correctly drive the probe heaters. If there is any uncertainty in this detection  
system the voltage can be forced to 220240 or 100127, and the frequency can be set to 50 or 60Hz.  
43  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
8.2.8 HEATER SSR FAULT CORRECTION  
In the event of one of the heater solid state relays (SSR) failing in single probe mode, the heater drive  
output can be swapped to the HEATER 2 output. Select “Heater 1<> Heater 2”.  
If you are running in dual probe mode but not using the calibration gas or purge solenoid outputs the  
solenoid outputs can be swapped to the 2 heater outputs. Select “Heaters <> CalPurge”. The default  
setting is “Normal”.  
8.2.9 SSR FAIL PROTECTION  
The transmitter continuously monitors the high voltage outputs on the solid state relays that power  
the probe and purge/cal solenoids. If it detects that any solenoids are not switching off correctly then  
it mechanically isolates all high voltage outputs in order to protect the instrument. This feature can be  
switched off by setting the SSR fail protection to disabled.  
44  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
9. ALARMS  
The 9060Z has 4 alarm relays and a built in alarm annunciator. If an alarm occurs, the ALARM light will  
flash. To find out what the alarm is, press the ALARM up key.  
When the ALARM up key has been pressed, the transmitter goes into the alarm display mode. In this  
mode some of the keys take on a special function.  
Key text  
RUN mode  
ALARM mode  
Return to RUN mode  
Next alarm  
SETUP / RUN  
*
*
*
DISPLAY / FUNCTION up  
DISPLAY / FUNCTION down  
ALARM / OPTION up  
ALARM / OPTION down  
ALARM ACCEPT / ENTER  
GAS 1 PURGE 1 / SENS IMP  
GAS 2 PURGE 2 / AUTO CAL  
Last alarm  
Enter ALARM display mode  
Enter ALARM log mode  
*
*
*
Alarm activated time  
Alarm acknowledged time  
Acknowledge alarm  
*
*
* This key is not used in the ALARM mode  
The common alarm relay is used to monitor faults within the transmitter and the probe. The other  
three alarm relays relate to the process gas. All relays have user defined actions.  
If one of the alarm events is removed from the common alarm using COMMISSIONING menu function  
#49, then this alarm event can be programmed for one of the other 3 alarm relays.  
When the alarm mode has been entered, the SETUP light flashes once a second until the SETUP key is  
pressed to return the transmitter to RUN mode.  
All relays have failsafe alarm contacts. That is –  
When the transmitter is turned off the contacts are open circuit  
When the transmitter is on but there are no alarms the contacts will be closed  
When there is a current unaccepted alarm event the contacts will be open circuit  
When there is a current accepted alarm event the state of the contacts will depend on the  
selection in the commissioning menu #50. See chapter 7.2.22, Operation of the alarm relay when  
an alarm is accepted.  
All alarms drive the alarm light on the front door.  
The light will be off if there are no alarms current  
The light will flash if there is a current alarm that has not been acknowledged  
The light will be on steady if there are current alarm(s) that have not been cleared  
The light will flash faster as more alarms occur  
9.1 COMMON ALARMS  
The events that drive the common alarm are –  
1. ‘Probe 1 Heater Fail’  
2. ‘Probe 2 Heater Fail’  
A heated probe (model 9060H) has been turned on for more than 20 minutes but is still not up to  
650°C (1200°F) or the probe temperature falls below 650°C.  
3. ‘Probe 1 High Impedance’  
45  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
4. ‘Probe 2 High Impedance’  
Oxygen probe or electrode failure. The sensor has a high impedance measurement. Replace the  
probe. This alarm is inhibited when the probes are less than 650°C (1200°F).  
5. ‘Probe 1 TC Open Circuit’  
6. ‘Probe 2 TC Open Circuit’  
Probe thermocouple is open circuit. The heater in heated probes will switch off.  
7. ‘Aux TC Open Circuit’  
The AUX thermocouple is open circuit. If the thermocouple is not needed, select “Disabled” in  
COMMISSIONING menu function #9, “Aux TC Type” or place a short circuit between terminals 7 & 8.  
8. ‘Ref Air Pump Fail’  
The reference air pump in the transmitter has been either unplugged or has gone open circuit  
(<20mA). Replace the reference air pump.  
9. ‘Ref Air Pump Overload  
The reference air pump in the transmitter has stalled and is drawing in excess of 300mA. The  
transmitter will attempt to restart the pump every minute for 1 second. Replace the reference air  
pump.  
10. ‘ADC Calibration Fail’  
The analogue to digital converter has been found to fall outside the normal calibration specifications.  
In this case the probe heater will automatically be turned off.  
11. ‘Alarm Log Fail’  
The alarm history is kept in an EEROM. This alarm will be raised if this memory device fails.  
12. ‘Output 1 Failure’  
13. ‘Output 2 Failure’  
The digital to analogue and voltage isolator circuit has been found to fall outside the normal  
calibration specifications. This check is only performed when the ‘AUTO CAL’ button is pressed. See  
chapter 10.1.2.  
14. ‘Heater 1 SSR Failure’  
15. ‘Heater 2 SSR Failure’  
One of the heater power control devices (SSR) has been found to have failed. See chapter 8.2.8,  
Heater SSR fault correction.  
16. ‘Heater SSR Leakage’  
One of the heater power control devices (SSR) has been found to have failed but the transmitter  
cannot determine which one(s) has failed.  
17. ‘Probe 1 Filter Blocked’  
18. ‘Probe 2 Filter Blocked’  
Blocked probe filter. This test is only performed when automatic purging of the probe is selected. See  
chapter 11. This alarm will not reset until the next purge cycle. The cycle can be initiated manually or  
automatically.  
46  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
19. ‘Gas 1 Calibration Error’  
20. ‘Gas 2 Calibration Error’  
At the end of an automatic gas calibration check cycle the oxygen reading was outside the values set  
in the COMMISSIONING menu #26 to 28 (#35 to 37). This alarm will not reset until the next calibration  
check cycle. The cycle can be initiated manually or automatically.  
21. ‘BBRAM Fail’  
All of the setup options are held in the battery backed memory (BBRAM). This is the battery shaped  
device at the bottom centre of the 17321 PCB labeled MEM1. This alarm will occur when this device  
fails and will need to be replaced.  
9.2 SELECTABLE PROCESS ALARMS  
The alarm relays 1, 2 and 3 are generally used to transmit a gas related alarm event. Any or all of the  
following functions can be selected for each relay.  
The trip levels and the delay times are set is in the COMMISSIONING menu.  
NOTE: The process alarms will only be activated if they are enabled in the SETUP function 6. See  
chapter 6.3.5.  
21. ‘Oxygen 1 Very Low’  
22. ‘Oxygen 2 Very Low’  
The measured oxygen level on the indicated probe has been below the trip level shown in  
COMMISSIONING menu function #38 for longer than the delay time shown in function #39.  
23. ‘Oxygen 1 Low’  
24. ‘Oxygen 2 Low’  
The measured oxygen level on the indicated probe has been below the trip level shown in  
COMMISSIONING menu function #40 for longer than the delay time shown in function #41.  
25. ‘Oxygen 1 High’  
26. ‘Oxygen 2 High’  
The measured oxygen level on the indicated probe has been above the trip level shown in  
COMMISSIONING menu function #42 for longer than the delay time shown in function #43.  
27. ‘Oxygen Deviation’  
The difference between the oxygen level measured on probe #1 and the oxygen level measured on  
probe #2 is greater than the trip level shown in COMMISSIONING menu function #44 for longer than  
the delay time shown in function #45.  
28. ‘Probe 1 Temperature Low’  
29. ‘Probe 2 Temperature Low’  
The probe temperature is under 650°C (1200°F). The oxygen reading is therefore invalid. If the probe  
heater has been on for more than 20 minutes and the temperature is less than 650°C (1200°F) a  
‘Probe # Heater Fail’ alarm will occur.  
NOTE: The ‘Probe # Temperature Low’ relay function is used with unheated probes to indicate oxygen  
reading is invalid (the probe is below 650°C (1200°F) ), in case the process temperature falls below this  
47  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
level. With heated probes this relay will be deenergised while the probe is heating up from ambient,  
making the contacts open circuit.  
30. ‘Cal 1 in Progress’  
31. ‘Cal 2 in Progress’  
A calibration check is occurring, either manual or automatic mode.  
32. ‘Purge 1 in Progress’  
33. ‘Purge 2 in Progress’  
A probe purge is occurring, either manual or automatic mode.  
9.3 ALARM RELAY OPTIONS  
The three process alarm relays (relays 1 to 3) are user defined. The relay action will depend on the  
selections made in the COMMISSIONING menu functions #45 to #48.  
The relay contacts are all designed to be “Fail Safe”. That is, they are  
open when the transmitter power is off  
closed when there is NOT an alarm condition  
open when an alarm occurs  
The contact will close circuit again (relay energised) when the alarm condition is acknowledged.  
Choose to have the relays react to any or all of the following alarms or warnings –  
Oxygen on probe 1 Low  
Oxygen on probe 2 Low  
Oxygen on probe 1 Very Low  
Oxygen on probe 2 Very Low  
Oxygen on probe 1 High  
Oxygen on probe 2 High  
Oxygen Deviation High  
Alarm  
Alarm  
Alarm  
Alarm  
Alarm  
Alarm  
Alarm  
Probe 1 Temperature Low  
Probe 2 Temperature Low  
Calibration check on probe 1 in  
Progress  
Warning  
Warning  
Warning  
Calibration check on probe 2 in  
Progress  
Warning  
Purge on probe 1 in Progress  
Purge on probe 2 in Progress  
Warning  
Warning  
In addition to the above process alarms that can activate the process alarm relays, any of the common  
alarm relay events that have been taken off the common alarm list in COMMISSIONING menu  
function #49 will appear on the process alarm relay lists in function #46 to #48.  
48  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
10. INSTRUMENT CALIBRATION  
10.1 CALIBRATION SUMMARY  
The 9060Z oxygen transmitter has a self calibration and diagnostic system built into the hardware and  
software. Once the reference voltages have been set into memory the self calibration system  
maintains the calibration of the transmitter. An automatic update of the zero and span is done every  
minute.  
The 420mA outputs can be automatically calibrated with the press of the AUTO CAL key or manually  
fine tuned.  
One entry for each probe offset is required to optimise the calibration of the oxygen measurement  
although an automatic system calibration check can be programmed into the transmitter using  
certified gasses.  
10.1.1 CALIBRATION OF THE INPUTS  
The calibration of the analogue inputs is based on the voltage of a temperature compensated voltage  
reference integrated circuit. There are 4 voltages generated from the standard reference voltage.  
They will vary by about 1% from one transmitter to another but can be trimmed by setting the actual  
voltages into CALIBRATION menu functions #1 to 4.  
The calibration should be done 30 minutes or more after the instrument has been on, approximately  
once every year. The calibration constants are retained in battery backed memory unless a ‘COLD  
START’ is performed. See chapter 10.2, Cold Start.  
Connect a 3 1/2 digit multimeter negative lead to the test point marked ‘COM’ in the centre of the  
17321 PCB (labeled ‘VREFS’). Measure the four voltages on the test point marked 1 to 4 with the  
positive lead of the multimeter. Enter the measured values in the CALIBRATION menu functions #1 to  
4. Whenever new values are entered, the D/A Section should be recalibrated. See chapter 10.1.2.  
10.1.2 CALIBRATION OF THE OUTPUTS  
The easiest way to calibrate the outputs is to select ‘Auto Calibrated’ in CALIBRATION menu functions  
#5 and 8. If this is selected the outputs will be directed away from the output terminals and back into  
an analogue input of the transmitter. The outputs are then tested and a zero and span calibration  
factor is recorded. If a more accurate calibration is required select ‘Manually Calibrated’ in  
CALIBRATION menu functions #5 and 8. This will inhibit the automatic calibration system overwriting  
the calibration factors. However, the zero and span factors will need to be manually set. To set the  
calibration factors use the following steps for each output:  
1. Select CALIBRATION menu function #5 (8) and set the option to ‘Calibrate 4mA’.  
2. Use the FUNCTION up key to go to function #6 (9).  
3. Set this to 4.00mA and press the ENTER key.  
4. Measure the actual output current with a digital multimeter.  
5. Use the OPTION keys to set the option to the value read on the digital multimeter and press  
ENTER.  
6. The output should go to 4.00mA. The value in function #6 (9) can be fine tuned if necessary.  
7. Go back to function #5 (8) and select ‘Calibrate 20mA’.  
8. Go up to function #7 (10) and set the option value to 20.00mA.  
9. Measure the actual output current with a digital multimeter.  
10. Use the OPTION keys to set the option to the value read on the digital multimeter and press  
ENTER.  
49  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
11. The output should go 20.00mA. The value in function #7 (10) can be fine tuned if necessary.  
12. Go back to function #5 (8) and select ‘Calibrate 4mA’ again and check that the output is still set  
to 4.00mA. If it is not exactly 4.00mA the value in function #6 (9) can be trimmed again.  
13. Go back to function #5 (8) and select ‘Manually Calibrated’.  
10.1.3 PROBE CALIBRATION  
There is only one calibration factor for the calibration of the oxygen probe. This is the sensor offset,  
and it is written on a tag attached to every probe.  
Use the FUNCTION keys to go to SETUP menu #1 (2) and then use the OPTION keys to set the value to  
the value written on the tag. If in doubt the best option is to set the offset to 0.0mV because this will  
only produce an error of around 0.1% oxygen in a combustion application for a 1mV error in the offset  
value.  
For information on checking the probe and transmitter system with a certified gas see chapter 11.  
10.2 COLD START  
The intention of the cold start function of the transmitter is to return all of the configuration set up  
back to the default values. There are 3 parts to the cold start:  
1. If a software upgrade is installed and the changes require a change to the menu functions then  
the transmitter will automatically do a cold start. The transmitter will also do a cold start when  
the power is turned on and the configuration / calibration factors are found to be corrupted. It  
will only reset either the configuration values or the calibration values if either of them is  
found to be corrupted.  
2. A cold start can be forced if the operator wants to reset the transmitter back to the default set  
up. See chapter 10.2.1, Forcing a Cold Start.  
3. Resetting the calibration factors back to the default values. See chapter 10.2.2, Resetting the  
Calibration Factors.  
10.2.1 FORCING A COLD START  
The cold start can be initiated by following these steps:  
1. Turn the transmitter power off.  
2. Use a fine point or screw driver to turn the COLD START switch on. The switch is at the bottom  
of the 17321 PCB, accessible through the window in the shield.  
3. Turn the transmitter back on.  
4. You will be prompted to select ‘Reset’ or ‘Cancel’ to the prompt ‘Reset Calibration Data?’.  
Press the DISPLAY up key to reset the calibration factors and the configuration data or the  
ALARM up key to only reset the configuration of the transmitter.  
5. Turn off the cold start switch when prompted by ‘Turn off C/Start Switch’.  
If the calibration factors have been reset, follow the instructions in chapter 10.1 to recalibrate the  
transmitter. If only the configuration has been reset it is also important to check items that will affect  
the transmitter outputs, number of probes, the serial communications, the solenoid configuration and  
other items.  
10.2.2 RESETTING THE CALIBRATION FACTORS  
If it is required to reset the calibration factors, follow the instructions in the previous section and  
select ‘Reset’ when prompted to ‘Reset Calibration Data?’.  
50  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
11. GAS CALIBRATION CHECK AND PURGE  
The oxygen sensor that is used in the Teledyne Analytical Instruments oxygen probe is extremely  
predictable, stable and reliable. For this reason, the calibration of a Teledyne Analytical Instruments  
oxygen system does not require the use of calibration gases. However, all Teledyne Analytical  
Instruments oxygen probes have a built in gas connection that does allow the accuracy of the probe to  
be checked insitu. This chapter describes the operation of this gas checking system.  
The 9060Z has a timer and solenoid driving system that can be configured to admit a certified  
calibration gas into the probe or an air supply to purge the probe filters through the gas connection.  
Both the calibration gas and the filter purge gas must be piped to the port on the probe labeled “CAL”.  
There are two solenoids drivers in the 9060Z transmitter. They can be used for any combinations of  
gas checking and probe filter purging. The only limitation is that if the transmitter is in dual probe  
mode, then solenoid #1 must be wired and plumbed to the probe #1 and solenoid #2 must be wired  
and plumbed to probe #2.  
The transmitter can also be configured to be in a MANUAL or AUTOMATIC purge and gas check mode.  
Function #  
20 (29)  
21 (30)  
22 (31)  
23 (32)  
24 (33)  
25 (34)  
26 (35)  
27 (36)  
28 (37)  
Function Description  
Options  
Solenoid 1 (2) Operation  
Solenoid 1 (2) Auto / Manual  
Solenoid 1 (2) Start Time  
Solenoid 1 (2) Period  
Calibration Gas / Purge  
Automatic / Manual  
00:00 to 23:45  
1 hour – 7 days  
1 – 90 seconds  
5 – 300 seconds  
0.1 – 20.9%  
Solenoid 1 (2) Duration  
Solenoid 1 (2) Post Freeze  
Oxygen Content Cal Gas 1 (2)  
Max Cal Gas 1 (2) Positive Error  
Max Cal Gas 1 (2) Negative Error  
0.1% to 3.0%  
0.1% to 3.0%  
Functions available if either Calibration Gas or Purge and Manual are selected  
Functions available if Purge and Automatic are selected  
Functions available if Calibration Gas and Automatic are selected  
11.1 PURGE  
For processes that have particulate present in the gas stream, probe filters are necessary and should  
be backpurged with sufficient frequency to avoid blocked filters. Select ‘Purge’ in function #20 (29)  
and ‘Automatic’ in function #30 and then set the times in function #22 (31) to #25 (34) The outputs  
can be frozen during purging if selected in COMMISSIONING menu function #19. A flow switch can be  
wired to terminals 16 and 17. If the flow is not sufficient at the end of the purge cycle to close the  
contacts of the flow switch a ‘Probe 1 (2) Filter Blocked’ alarm is generated. If a flow switch is not  
being used a short circuit should be placed across terminals 16 and 17 to suppress the ‘Probe 1 (2)  
Filter Blocked’ alarm.  
11.2 CALIBRATION GAS  
If it is required to automatically check the oxygen probe and transmitter on line, select ‘Calibration  
Gas’ in function #20 (29) and ‘Automatic’ in function #30 and then set the times in function #22 (31)  
to #25 (34) and the gas levels in function #26 (35) to 28 (37). If the transmitter measurement does not  
fall within the limits set in function #26 (35) to 28 (37) a ‘Gas1 (2) Calibration Error’ alarm will be  
generated. If the transmitter has ‘Automatic’ selected in function #21 (30) and one of the GAS/PURGE  
keys is pressed in RUN mode, the automatic cycle will be started and the alarms will be checked. The  
cycle can be stopped at any time by pressing the same key during the cycle.  
51  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
12. SOFTWARE UPGRADES  
The software for the 9060Z series transmitters is contained in memory inside the microprocessor  
integrated circuit (IC). The transmitter therefore does not require an external memory IC for the  
software. Provision has been made to enable the software to be upgraded in the field without the use  
of a computer. If an upgrade is to be made to the software it will be supplied in a 32 pin EEROM IC  
type 29F010B. To do the upgrade, use the following steps:  
1. Turn the power off to the transmitter at the power switch  
2. Remove the hose from the reference air pump (if fitted)  
3. Unplug the reference air DC power lead (if fitted)  
4. Undo the two M4 screws at the top and bottom of the main shield  
5. Remove the earth screw from the main shield on the right hand side of the cabinet  
6. Lift out the main shield around the cable glands  
7. Fold out the two wings of the blue socket labeled “FIRMWARE UPGRADE SOCKET”. Plug the  
29F010B into the socket. Carefully note the direction of the IC. Pin 1 is identified on the PCB and  
pin 1 on the IC has a small round indentation next to the pin.  
8. Hold down the DISPLAY down and the ALARM LOG keys and then turn on the power  
9. Release the keys when the message “Verifying the EEROM” will be shown. The transmitter is  
confirming that the new firmware is valid.  
10. After the EEROM has been read and confirmed this message will be shown:  
Upgrade Firmware?  
Model 9060Z  
New Version: 1.##  
Upgrade  
Cancel  
11. Press the DISPLAY up key to ‘Upgrade’ the firmware.  
12. Press the ALARM up key to cancel the upgrade and continue the start up of the current firmware.  
13. When the transmitter has upgraded the firmware it will start up the normal initialization. The  
version of the firmware is shown on the startup screen.  
Upgrading Firmware.  
Analyser will reset itself in approx. 15 seconds.  
Do not turn off the power  
NOTE: While the firmware is being upgraded it is essential that the power is not  
turned off otherwise the programme will be corrupted and will not be recoverable.  
14. Turn the transmitter power off.  
15. Unplug the EEROM by pressing out the wings of the blue upgrade socket. Keep the EEROM in a  
safe place. It can be used to upgrade any number of transmitters as required.  
16. Replace the main shield and secure it with the three screws.  
17. Reconnect the reference air pump hose and plug (if fitted).  
18. Turn the power back on.  
NOTE: The transmitter may do an automatic COLD START after the upgrade. The words  
‘Cold Start’ will be shown on the display in step 14 if the cold start has been performed.  
The calibration will NOT be changed but the configuration may have been changed.  
52  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
13. TROUBLESHOOTING  
The Teledyne Analytical Instruments oxygen transmitter has proved an extremely reliable instrument.  
The trouble shooting guide here is based on an analysis of the potential problem that may occur after  
many years of operation in the field. A current list of problems and solutions can be found on the  
Teledyne Analytical Instruments web site at www.teledyneai.com when any are identified.  
13.1 FIRST APPROACH  
The probe will not heat up.  
Turn the power off, remove and check 2 x 20mm glass fuses. If either of them is blown, check the  
probe heater resistance between the two white wires that come from the probe. It should be 110 +/‐  
15 ohms, and both wires should be open circuit to earth. If not, replace the probe.  
Is the ‘B’ shown on the bottom left hand corner of the display? If not, check that the burner is enabled  
with either a link between terminals 10&11, BURNER INPUT, or is enabled with a connection to a  
voltagefree contact from the main fuel valve to these terminals.  
The display is blank and there is no backlight on.  
Measure the power supply voltages at the test points labeled COMM and +5v at the right hand side of  
the main PCB 17321. It should be 5 +/0.15v.  
Turn off the power and remove the main shield. Check that the plugin fuse FS1 has not blown. It  
should only be replaced with a 1A, 250v fuse if it has blown.  
Is the mains power being supplied to terminals 36&37 and is it between 100 and 240VAC? If may be  
necessary to replace the switch mode power supply, PS5, APC5S.  
An ‘ADC Calibration Fail’ alarm has occurred.  
Turn the transmitter off and remove the main shield. Turn the transmitter back on and measure the  
voltages on the ‘Acom’ test points. With the negative lead on the centre terminal the other 2  
terminals should measure 12 +/0.3v. If not, replace the small DCDC converter PS4.  
An ‘Output 1 (2) Failure’ alarm has occurred.  
Turn the transmitter off and remove the main shield. Turn the transmitter back on and measure the  
voltages on the ‘D1com’ test points. With the negative lead on the centre terminal the other 2  
terminals should measure 12 +/0.3v. If not, replace the small DCDC converter PS4.  
A ‘Heater 1 (2) Failure’ alarm has occurred.  
The software has found that the SSR4 has failed. If only one probe is being used and you need to gets  
the transmitter back working quickly, use the CALIBRATION menu function 18, Heater SSR Select, to  
change the heater output terminals. See chapter 8.2.8, Heater SSR fault correction.  
A ‘BBRAM Fail’ alarm has occurred  
Replace the BBRAM, MEM1 on the 17321 main PCB.  
13.2 DETAILED FAULT ANALYSIS  
The 9060Z transmitter has a diagnostics mode built into the software. This mode allows detailed  
analysis of the hardware of the transmitter, but does require a level of competence in electronics.  
The diagnostics mode is selectable by turning the ‘Diagnostics’ DIP switch to ON and then turn on the  
power. There is a separate Diagnostics Mode Manual available that describes its use. Ask Teledyne  
Analytical Instruments for more details.  
53  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
14. MAINTENANCE  
14.1 TRANSMITTER MAINTENANCE  
The 9060Z transmitter has several hardware checking systems that confirm the correct operation of  
the transmitter and raise an alarm if there is a fault detected. Because of this system, the transmitter  
only requires periodic maintenance. Teledyne Analytical Instruments recommends an annual check of  
the transmitter.  
The annual check should include the following items –  
1. Measure and record the reference voltages, and check that the correct voltages are set in the  
calibration menu  
2. Check that the reference air flow from the port on the bottom of the transmitter case is between  
100 and 300cc/m  
3. Check for contamination and potential blockage of the reference air filter that is mounted on the  
bottom of the transmitter case  
4. If an external burner interlock has been connected (terminals 10 & 11), check that if one of the  
wires is disconnected the probe heater is disabled (9060H probes only)  
5. Check the integrity of the cable. Particularly check the main power supply cable and the probe  
cable for any damage.  
6. Check the integrity of the earth connections in the transmitter  
14.2 CLEANING  
The 9060Z transmitter is made in an IP65 / IP54 rated diecast case. It is painted with a tough ripple  
finish power coating to maintain a clean look. However, if it is necessary to clean the case use either a  
warm soapy water solution or an isopropyl alcohol.  
A build up of dust on or around the transmitter should be removed before the cabinet door is opened.  
If there is a buildup of dust inside the cabinet check the door oring seal for damage and replace it if  
necessary.  
Take care when wiping the display window to avoid scratching the surface.  
If the filters of the probe (if fitted) become blocked, use an ultrasonic cleaner with a few drops of  
household detergent. If the filters cannot be cleaned this way replace the sintered filters.  
The filter in the base of the analyzer may become blocked if the analyzer is operating in a dusty  
environment. It is available as a field replacement item.  
14.3 REPLACEMENT PARTS  
The following list identifies field replaceable parts –  
Stock number  
Description  
18020  
18021  
15039  
14105  
14093  
18064  
11040  
11041  
14123  
17321 Main PCB  
17322 Display PCB  
MP24E Reference air pump  
Fuse M205 3.15A 250V fuse (pack of 5)  
Instruction manual  
EEROM 9060x firmware  
FIL230 Titanium sintered filter, 30um  
FIL215 Titanium sintered filter, 15um  
Filter – analyzer bulkhead  
54  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
15. INDEX  
9060H Oxygen Probes ....................................................................................................................... 9  
Alarm connections .......................................................................................................................... 23  
Alarm log clearing ........................................................................................................................... 43  
Alarm relay options ........................................................................................................................ 50  
Alarms.............................................................................................................................................. 47  
Alarms, common.............................................................................................................................. 47  
Alarms, process ............................................................................................................................... 49  
Alarms, process enabling ................................................................................................................ 35  
Auxiliary Thermocouple ............................................................................................................ 17, 23  
Calibration........................................................................................................................................ 51  
Calibration gas check ...................................................................................................................... 27  
Calibration Mode............................................................................................................................. 45  
Calibration of the sensor.................................................................................................................. 27  
Cold start ......................................................................................................................................... 52  
Commissioning Mode ...................................................................................................................... 37  
Common Alarms .............................................................................................................................. 23  
Communications port options.......................................................................................................... 42  
Connecting the power...................................................................................................................... 27  
Connections, electrical..................................................................................................................... 19  
Damping factor ................................................................................................................................ 35  
Date / Time ...................................................................................................................................... 39  
Display units, oxygen........................................................................................................................ 30  
Dust in the flue gas .......................................................................................................................... 28  
Filter purging.................................................................................................................................... 27  
Fuel selection.................................................................................................................................... 42  
Gas calibration check ....................................................................................................................... 53  
Graphical display .............................................................................................................................. 29  
Heater interlock relays ..................................................................................................................... 21  
Installing a 9060H Oxygen Probe ..................................................................................................... 15  
Keypad............................................................................................................................................... 31  
Lower line ......................................................................................................................................... 35  
Mains voltage detection ................................................................................................................... 46  
MODBUS connection......................................................................................................................... 25  
Mounting the Transmitter................................................................................................................. 15  
Number of probes ............................................................................................................................ 39  
Outputs, 420mA......................................................................................................................... 37, 39  
Pressure compensation..................................................................................................................... 40  
Pressure Transducer.......................................................................................................................... 28  
Probe Cable....................................................................................................................................... 22  
55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
Probe offset entry ............................................................................................................................ 34  
Purge ................................................................................................................................................ 53  
Purge and calibration connections ................................................................................................... 24  
Reference air connection.................................................................................................................. 25  
Reference air pump options.............................................................................................................. 42  
Reference Voltages........................................................................................................................... 46  
Run mode.......................................................................................................................................... 27  
Setup Mode ...................................................................................................................................... 33  
Setup mode functions ...................................................................................................................... 34  
Software upgrades............................................................................................................................ 55  
SPECIFICATIONS................................................................................................................................ 12  
Specifications Hardware.................................................................................................................. 11  
Specifications Software ................................................................................................................... 14  
SSR Fault Correction.......................................................................................................................... 46  
Stratification ..................................................................................................................................... 28  
Temperature units............................................................................................................................ 40  
Troubleshooting ......................................................................................................................... 57, 58  
Voltage, Mains Supply....................................................................................................................... 11  
Warnings............................................................................................................................................ 7  
56  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
APPENDIX 1, CONSTITUENT VALUES FOR VARIOUS FUELS  
If the transmitter is set up to have readout or output of combustibles, oxygen deficiency, maximum  
carbon dioxide or combustion efficiency, then the fuel constituents must be entered. Any or all of the  
variables can be modified and entered in COMMISSIONING menu functions #50 to 56. (See chapter  
7.2.23). Your fuel supplier or chemist should be able to give you all these values.  
A
H
O
N
S
is the heat of combustion of the fuel per gram atom of contained carbon.  
is the H/C atom ratio in the fuel.  
is the O/C atom ratio in the fuel.  
is the N/C atom ratio in the fuel.  
is the S/C atom ratio in the fuel.  
M is the ratio of H2O molecules to C atoms in the fuel  
FUEL  
A
H
O
N
S
M
Blast Furnace Gas  
Coke Oven Gas  
Producer Gas  
50.55  
256.88  
101.98  
209.9  
176.4  
166.1  
172.59  
157.58  
149.65  
145.18  
145.54  
142.25  
136.52  
110.91  
116.88  
104.98  
99.63  
0.08  
5.6  
1.3 3.08 b  
0.25 0.23 b  
a
a
a
1.18  
3.86  
2.69  
2.34  
3.97  
2.01  
1.83  
1.71 ‐  
1.6 ‐  
1.44 ‐  
1.25  
2.26  
0.74  
0.35  
0.11  
1.02  
2.9 b  
Natural Gas  
0
0
0
0.1  
0
0
0
0
0
0
0
0
Propane, Natural  
Butane, Refinery  
Methanol  
Gasoline, Motor  
No 1 Distillate Oil  
No 2 Distillate Oil  
No 4 Fuel Oil  
No 5 Residual Oil  
No 6 Residual Oil  
Wood, NonResinous  
Coal, Bituminous  
Coal, Anthracite  
Coke  
1 ‐  
0
0
0
0
0 ‐  
0 ‐  
0 ‐  
0.01  
0
0
0
0
0
0
0
0
0
0.01  
0
1.07  
0.05  
0.05  
0 c  
0
0.03  
0.01 0.04  
0.01  
0.01 0.01  
0
a.  
b.  
c.  
The moisture level varies depending on the process details. The calculated values  
assume M = O.  
The sulfur level varies depending on the process details. The calculated values  
assume S = O.  
Variable.  
Values calculated from the North American Combustion Handbook, Tables 2.1a and 2.1b.  
57  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
APPENDIX 2, PROBE EMF TABLES  
ZIRCONIA OXYGEN SENSOR OUTPUT (mV) PROBE TYPE 9060H, SENSOR  
OXYGEN  
%
Probe EMF  
@ 720°C (1320°F)  
OXYGEN  
%
Probe EMF  
@ 720°C (1320°F)  
20.95  
0
10  
15.93  
20.5  
20  
19.5  
19  
18.5  
18  
17.5  
17  
16.5  
16  
15.5  
15  
14.5  
14  
0.517  
1.046  
1.589  
2.147  
2.719  
3.306  
3.911  
4.533  
5.173  
5.834  
6.515  
7.219  
7.947  
8.7  
9.5  
9
8.5  
8
7.5  
7
6.5  
6
5.5  
5
4.5  
4
3.5  
3
17.033  
18.196  
19.426  
20.73  
22.12  
23.607  
25.204  
26.93  
28.808  
30.867  
33.145  
35.695  
38.59  
41.94  
13.5  
13  
12.5  
12  
11.5  
11  
10.5  
9.481  
10.292  
11.134  
12.011  
12.925  
13.881  
14.881  
2.5  
2
45.913  
50.797  
57.135  
66.182  
82.168  
99.518  
114.347  
1.5  
0.9  
0.4  
0.2  
0.1  
'K' Type TC 29.965 mV @ 720°C (1320°F)  
These tables are based on the Nernst equation: O2  
ꢆ₂  
ꢇꢈ.ꢉꢊ  
ꢀ ꢁꢂꢃ ꢅ  
EMF  
=
Where T is temperature° K  
ꢌꢍꢎ.ꢍꢏꢐ  
58  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
ZIRCONIA OXYGEN PROBE OUTPUT (mV) PROBE TYPE 9060UL/UH  
These tables are based on the Nernst equation:  
59  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
ꢆ₂  
ꢇꢈ.ꢉꢊ  
ꢀ ꢁꢂꢃ ꢅ  
EMF  
=
Where T is temperature° K  
ꢌꢍꢎ.ꢍꢏꢐ  
Thermocouple information from NIST ITS90 data tables  
APPENDIX 3, % OXYGEN SCALE TO LOGARITHMIC  
OXYGEN  
%
FULL SCALE  
%
0.1  
0.15  
0.2  
0.3  
0.4  
0.6  
0.8  
1
1.5  
2
3
4
6
0
7.65  
13.1  
20.7  
26.2  
33.8  
39.2  
43.5  
51.1  
56.5  
64.2  
69.6  
77.3  
82.7  
86.9  
90.4  
93.3  
95.8  
98  
8
10  
12  
14  
16  
18  
20  
100  
60  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
APPENDIX 4, MODBUS™  
The Teledyne Analytical Instruments 9060Z transmitter has the ability to work as a Modbus slave node  
on serial RS232 or 3wire RS485 via RTU mode transmissions. By accessing information stored in the  
input and holding registers it is possible to access runtime variables, alarm conditions and modify the  
device configuration.  
The implementation of the Modbus protocol is limited to the specific command set relevant to  
reading and writing register variables. Attempting to send unrecognised commands will result in  
appropriate error responses  
MODBUS™ Functions Supported:‐  
0x03: Read Holding Registers  
0x04: Read Input Registers  
0x06: Write Single Holding Register  
0x08: (return query data – for loopback testing)  
0x10: Write Multiple Holding Registers  
0x16: Mask Write Holding Register  
0x17: Read/Write Multiple Holding Registers  
0x2B: Encapsulated Interface (read device  
information)  
0x41: Special Instruction Function  
Serial Configuration  
The serial configuration for the slave device is accessed in the Commissioning menu of the  
transmitter. Default settings are highlighted in bold.  
Baud Rate  
Parity  
Stop Bits  
Interface  
9600, 19200, 38400, 57600  
Even, Odd, None  
1
RS485, RS232  
All holding register addresses contain a single 16 bit value, however some variables span multiple  
holding registers to yield a single 32 bit variable.  
When reading register values be aware that the byte order of data within registers may differ from  
that of the target system.  
Modbus Protocol  
The modbus over serial line protocol defines a messaging system for master/slave communications.  
Only one master device may be connected to any network with one or several slave nodes.  
In unicast mode, the master initiates communication with a specific slave node by sending a request  
message. The slave processes the request and returns a reply message containing the requested  
information or a confirmation that the request was fulfilled.  
Modbus requests are transmitted as a formed request frame with a CRC for data integrity checking.  
The frame for each request and reply type will vary somewhat, but will always have the basic  
structure described below:  
1 byte  
1 byte  
up to 255 bytes  
2 bytes  
slave address  
function code  
data  
checksum  
Internal Representation of Dates & Alarm Status  
The transmitter stores all dates as an unsigned 32bit count of seconds elapsed since 1Jan 2004.  
61  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
Alarm status is stored as an integer value:  
0
clear  
2
3
active  
1
self cleared  
acknowledged  
9060Z Transmitter Modbus Input Register Table for firmware v1.16 and newer  
62  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
63  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
64  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
9060Z Transmitter Modbus Holding Register Table for firmware v1.16 and newer  
65  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
66  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 9060Z Manual  
0x41 Special Instruction Function  
This command allows you to interact with the transmitter allowing access to functionality that would  
otherwise only be accessible via the keypad. The request length will vary from one command to the  
next depending on what arguments are required. Requests are formed the same as other modbus  
requests:  
1 byte  
1 byte  
1 byte  
0 4 bytes  
2 bytes  
slave address  
command code  
function code arguments  
checksum  
Function Code Command  
Arguments  
0x01  
0x02  
0x03  
0x04  
0x05  
Accept All Active Alarms  
remote set key down  
remote reset key down  
Initiate Probe Impedance Check none  
Set Internal Clock 4 byte rtc*  
none  
1 byte keymask  
1 byte keymask  
* date expressed as seconds since 1Jan 2004  
67  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Tech Craft TV Video Accessories NTR60 User Manual
Technicolor Thomson DVD Recorder DTH8555 X User Manual
Toshiba Graphics Tablet WT8 A Series User Manual
Toshiba Marine Radio P850 P855 User Manual
TRENDnet Router 201 User Manual
Victory Refrigeration Refrigerator ACRS ID S STS User Manual
ViewSonic Projection Television VS11529 User Manual
Vivotek Digital Camera IP7139 User Manual
Weber Gas Grill 55263 User Manual
Weed Eater Lawn Mower 177019 User Manual