Teledyne Oxygen Equipment 3010MB User Manual

OPERATING INSTRUCTIONS  
Model 3010MB  
Percent Paramagnetic  
Oxygen Analyzer  
Bulkhead Mount Control Unit, PN D-71385  
Bulkhead Mount Analysis Unit, PN D-71394  
DANGER  
HIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS MONITORING  
SYSTEM.  
PERSONAL PROTECTIVE EQUIPMENT MAY BE REQUIRED WHEN SERVICING THIS SYSTEM.  
HAZARDOUS VOLTAGES EXIST ON CERTAIN COMPONENTS INTERNALLY WHICH MAY PERSIST  
FOR A TIME EVEN AFTER THE POWER IS TURNED OFF AND DISCONNECTED.  
P/N xxxxxx  
12/21/99  
ECO:#99-0000  
ONLYAUTHORIZEDPERSONNELSHOULDCONDUCTMAINTENANCEAND/ORSERVICING. BEFORE  
CONDUCTING ANY MAINTENANCE OR SERVICING CONSULT WITH AUTHORIZED SUPERVISOR/  
MANAGER.  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Paramagnetic Oxygen Analyzer  
Table of Contents  
Specific Model Information................................ iv  
Preface................................................................v  
Part I: Control Unit, Model MB..............Part I: 1-1  
Part II: Analysis Unit, Model M ............Part II: 1-1  
Appendix......................................................... A-1  
Teledyne Analytical Instruments  
iii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 3010MB  
SpecificModelInformation  
The instrument for which this manual was supplied may incorporate  
one or more options not supplied in the standard instrument. Commonly  
available options are listed below, with check boxes. Any that are incorpo-  
rated in the instrument for which this manual was supplied are indicated by a  
check mark in the box.  
Instrument Serial Number: __________________________  
.
iv  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Paramagnetic Oxygen Analyzer  
Preface  
Overview  
The Analytical Instruments Model 3010MB Percent Paramagnetic  
Oxygen Analyzer is a versatile microprocessor-based instrument for detect-  
ing parts-per-million (ppm) levels of oxygen in a variety of background  
gases. It is a “split architecture” instrument. This means that a general pur-  
pose Control Unit, designed for nonhazardous areas only, remotely controls  
a specially designed Analysis Unit, or remote probe.  
Part I of this manual covers the Model 3010MB General Purpose  
NEMA 4 Bulkhead mount Control Unit only. This Control Unit is for  
outdoor/indoor use in a nonhazardous environment. The Analysis Units (or  
Remote Probes) it controls, can be designed for a variety of hazardous  
environments. Part II of this manual covers the 3010M Analysis Unit.  
Typical Applications  
A few typical applications of the Model 3010MB are:  
Monitoringinertgasblanketing  
Airseparationandliquefaction  
Chemicalreactionmonitoring  
Semiconductormanufacturing  
Petrochemicalprocesscontrol  
Qualityassurance  
Gasanalysiscertification.  
Teledyne Analytical Instruments  
v
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 3010MB  
Main Features of the Analyzer  
The Model 3010MB Percent Oxygen Analyzer is sophisticated yet  
simple to use. The main features of the analyzer include:  
A 2-line alphanumeric display screen, driven by microprocessor  
electronics, that continuously prompts and informs the operator.  
High resolution, accurate readings of oxygen content from low %  
levels through 100%. Large, bright, meter readout.  
Stainlesssteelsamplesystem.  
Versatile analysis over a wide range of applications.  
Microprocessorbasedelectronics:8-bitCMOSmicroprocessor  
with 32 kB RAM and 128 kB ROM.  
Three user definable output ranges (from 0-5 % through 0-100  
%) allow best match to users process and equipment.  
Auto Ranging allows analyzer to automatically select the proper  
preset range for a given measurement. Manual override allows  
the user to lock onto a specific range of interest.  
Twoadjustableconcentrationalarmsandasystemfailurealarm.  
Extensive self-diagnostic testing, at startup and on demand, with  
continuouspower-supplymonitoring.  
RS-232 serial digital port for use with a computer or other digital  
communicationdevice.  
Analog outputs for concentration and range identification.  
(0-1 VDC standard, and isolated 4–20 mADC)  
vi  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
OPERATING INSTRUCTIONS  
Model 3010MB  
Oxygen Analyzer  
Part I: Control Unit  
NEMA 4 Bulkhead Mount  
Part Numbers: D-71385  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 3010MB Percent Paramagnetic Oxygen Analyzer  
Table of Contents  
1 Introduction  
1.1 Overview........................................................................ 1-1  
1.2 Control Unit Inner Control Panel .................................... 1-1  
1.3 Recognizing Difference Between LCD & VFD............... 1-3  
1.4 Control Unit Interface Panel ........................................... 1-4  
2 OperationalTheory  
2.1 Introduction .................................................................... 2-1  
2.2 Electronics and Signal Processing ................................ 2-1  
3 Installation  
3.1 Unpacking the Control Unit............................................ 3-1  
3.2 Mounting the Control Unit .............................................. 3-1  
3.3 Electrical Connections ................................................... 3-3  
3.4 Testing the System......................................................... 3-12  
4 Operation  
4.1 Introduction .................................................................... 4-1  
4.2 Using the Data Entry and Function Buttons ................... 4-2  
4.3 The System Function ..................................................... 4-3  
4.3.1 Tracking the O2 Readings during Calibration ......... 4-4  
4.3.2 Setting up an Auto-Cal........................................... 4-5  
4.3.3 Password Protection .............................................. 4-5  
4.3.3.1 Entering the Password ................................... 4-6  
4.3.3.2 Installing or Changing the Password ............. 4-7  
4.3.4 Logout.................................................................... 4-8  
4.3.5 System Self-Diagnostic Test .................................. 4-9  
4.3.6 Version Screen ...................................................... 4-10  
4.3.7 Showing Negative Oxygen Readings .................... 4-10  
4.4 The Zero and Span Functions ....................................... 4-10  
4.4.1 Zero Cal ................................................................. 4-11  
ii: Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Part I: Control Unit  
4.4.1.1 Auto Mode Zeroing ........................................ 4-11  
4.4.1.2 Manual Mode Zeroing.................................... 4-12  
4.4.1.3 Cell Failure .................................................... 4-13  
4.4.2 Span Cal................................................................ 4-14  
4.4.2.1 Auto Mode Spanning ..................................... 4-14  
4.4.2.2 Manual Mode Spanning................................. 4-16  
4.4.3 Span Failure .......................................................... 4-16  
4.5 The Alarms Function...................................................... 4-15  
4.6 The Range Function ...................................................... 4-18  
4.6.1 Setting the Analog Output Ranges......................... 4-19  
4.6.2 Fixed Range Analysis............................................ 4-20  
4.7 The Analyze Function .................................................... 4-20  
4.8 Signal Output ................................................................. 4-21  
5 Maintenance  
5.1 Fuse Replacement......................................................... 5-1  
5.2 System Self Diagnostic Test........................................... 5-2  
5.3 Major Internal Components............................................ 5-3  
A Appendix  
Model 3010MB Specifications ............................................... A-1  
Teledyne Analytical Instruments  
Part I: iii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 3010MB Percent Paramagnetic Oxygen Analyzer  
iv: Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Paramagnetic Oxygen Analyzer  
Part I: Control Unit  
Introduction  
1.1 Overview  
The Analytical Instruments Model 3010MB Analyzer Control Unit,  
together with a 3010M Analysis Unit, is a versatile microprocessor-based  
instrument for detecting percent amounts of oxygen in a variety of gases.  
Part I, this part, of this manual covers the Model 3010MB series Gen-  
eral Purpose NEMA 4 Bulkhead mount Control Units. (The Analysis Unit is  
covered in Part II of this manual.) The Control Unit and Analysis Unit is for  
indoor/outdoor use in a nonhazardous environment only.  
1.2 Control Unit Inner Control Panel  
The standard 3010MB Control Unit is housed in a rugged NEMA 4  
metal case with all remote controls and displays accessible from the inner  
control panel. See Figure 1-1. The inner control panel has a digital meter, an  
alphanumeric display, and thirteen buttons for operating the analyzer.  
Part I: 1-1  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1 Introduction  
Model 3010MB  
Outer Door  
(Open)  
Teledyne Analytical Instrument  
Viewing  
Window  
0.0  
AL-1  
%
Anlz  
Outer Door  
Latch  
3010MB Series  
Paramagnetic Oxygen Analyze  
Figure 1-1: Front of Unmounted Control Unit  
Function Keys: Six touch-sensitive membrane switches are used to  
change the specific function performed by the analyzer:  
Analyze Perform analysis for oxygen content of a sample gas.  
System Performsystem-relatedtasks(describedindetailin  
chapter 4, Operation.).  
Span  
Zero  
Span calibrate the analyzer.  
Zero calibrate the analyzer.  
Alarms Set the alarm setpoints and attributes.  
Range  
Set up the 3 user definable ranges for the instrument.  
1-2: Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Paramagnetic Oxygen Analyzer  
Part I: Control Unit  
Data Entry Keys: Six touch-sensitive membrane switches are used to  
input data to the instrument via the alphanumeric VFD display:  
Left & Right Arrows  
Selectbetweenfunctionscurrently  
displayed on the VFD screen.  
Up & Down Arrows  
Increment or decrement values of  
functionscurrentlydisplayed.  
Enter Moves VFD display on to the next screen in a series. If  
none remains, returns to the Analyze screen.  
Escape Moves VFD display back to the previous screen in a  
series. If none remains, returns to the Analyze screen.  
Digital Meter Display: The meter display is a LED device that  
produces large, bright, 7-segment numbers that are legible in any lighting. It  
is accurate across all analysis ranges. The –MB model produce continuous  
readout from 0-1% through 0-100%.  
Alphanumeric Interface Screen: The backlit VFD screen is an easy-  
to-use interface between operator and analyzer. It displays values, options,  
and messages that give the operator immediate feedback.  
I/O Power Button: The red I/O button switches the instrument power  
between I (ON) and O (a Keep-Alive state). In the O state, the instrument’s  
circuitry is operating, but there are no displays or outputs.  
CAUTION: The power must be disconnected to fully  
disconnect power from the instrument. When  
chassis is exposed or when access door is open  
and power cable is connected, use extra care to  
avoid contact with live electrical circuits .  
Access Door: For access to the electronics and interface panel, the front  
panel swings open when the latch in the upper right corner of the panel is  
pressed all the way in with a narrow gauge tool. Accessing the main circuit  
board and other electronics requires unfastening the rear panel screws and  
sliding the unit out of the case.  
1.3 Recognizing Difference Between LCD &  
VFD  
LCD has GREEN background with BLACK characters. VFD has  
DARK background with GREEN characters. In the case of VFD - NO  
CONTRAST ADJUSTMENT IS NEEDED.  
Part I: 1-3  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1 Introduction  
Model 3010MB  
1.4 Control Unit Interface Panel  
The Control Unit interface panel, shown in Figure 1-2, contains the  
electrical terminal blocks for external inputs and outputs. The input/output  
functions are described briefly here and in detail in the Installation chapter of  
thismanual.  
Figure 1-2: Model 3010MB Rear Panel  
1-4: Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Paramagnetic Oxygen Analyzer  
Part I: Control Unit  
Power Connection  
Analog Outputs  
AC power source, 100-240VAC, 50/60  
Hz  
0-1 V dc concentration and 0-1 V dc  
range ID. Isolated 4-20 mA dc and 4-20  
mA dc range ID.  
Alarm Connections 2 concentration alarms and 1 system  
alarm.  
RS-232 Port  
Serialdigitalconcentrationsignaloutput  
and control input.  
Remote Probe  
Remote Span/Zero  
Providesallelectricalinterconnecttothe  
Analysis Unit or Remote Probe.  
Digital inputs allow external control of  
analyzercalibration.  
Calibration Contact To notify external equipment that  
instrument is being calibrated and  
readings are not monitoring sample.  
Range ID Contacts  
Remote Probe  
Four separate, dedicated, range relay  
contacts. Low, Medium, High, Cal.  
Interfaces with an Analysis Unit or  
RemoteProbe(externalsensor/sample  
system).  
Network I/O  
Serialdigitalcommunicationsforlocal  
network access. For future expansion.  
Not implemented at this printing.  
Note: If you require highly accurate Auto-Cal timing, use external  
Auto-Cal control where possible. The internal clock in the  
Model 3010MB is accurate to 2-3 %. Accordingly, internally  
scheduled calibrations can vary 2-3 % per day.  
Part I: 1-5  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1 Introduction  
Model 3010MB  
1-6: Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Paramagnetic Oxygen Analyzer  
Part I: Control Unit  
Operational Theory  
2.1 Introduction  
The Model 3010MB Oxygen Analyzer Control Unit uses an 80C31  
microcontroller with 32 kB of RAM and 128 kB of ROM to control all  
signal processing, input/output, and display functions for the Model 3010MB  
analyzer. (The sample system and Paramagnetic sensor are covered in Part  
II, Analysis Unit, in this manual.) System power is supplied from a universal  
power supply module designed to be compatible with any international  
power source.  
2.2 Electronics and Signal Processing  
All of the Analyzer electronics are located on Printed Circuit Board  
(PCB) assemblies inside the Control Unit chassis. The PCB locations are  
illustratedinsection5,Maintenance.  
Refer to Figure 2-1, Block Diagram of the 3010MB CU Electronics:  
In the presence of oxygen, the sensor (in the Analysis Unit) generates a  
current. A current to voltage amplifier (in the Control Unit) converts this  
current to a voltage.  
The second stage amplifier amplifies the voltage. The output from the  
second stage amplifier is sent to an 18-bit analog to digital converter con-  
trolledbythemicroprocessor.  
The digital concentration signal—along with input from the control  
panel—is processed by the microprocessor, and appropriate control signals  
are directed to the display, alarms and communications port as well as to the  
gas control valves in the Analysis Unit.  
The same digital information is also sent to a 12 bit digital to analog  
converter that produces the 4-20 mA dc and the 0-1 V dc analog concentra-  
tion signal outputs, and the analog range ID outputs.  
The microprocessor monitors the power supply, and activates the  
system failure alarm if a malfunction is detected.  
Teledyne Analytical Instruments  
Part I: 2-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2 Operational Theory  
Model 3010MB  
Second  
Stage  
Current  
to Voltage  
Amplifier  
Amplifier  
Sensor*  
A t  
Conv  
*From  
Analysis  
Unit  
Power  
Supply  
Auto  
Rang  
System  
Failure  
Alarm  
M
Pro  
To Analysis  
Unit  
Control Valves  
Display  
Self Test  
Signal  
Figure 2-1: Block Diagram of the 3010TB CU Electronics  
2-2: Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Paramagnetic Oxygen Analyzer  
Part I: Control Unit  
Installation  
Installation of Model 3010MB Analyzers includes:  
1. Unpacking, mounting, and interconnecting the Control Unit and  
the Analysis Unit  
2. Making gas connections to the system  
3. Making electrical connections to the system  
4. Testing the system.  
This chapter covers installation of the Control Unit. (Installation of the  
Analysis Unit is covered in Part II of this manual.)  
3.1 Unpacking the Control Unit  
The analyzer is shipped with all the materials you need to install and  
prepare the system for operation. Carefully unpack the Control Unit and  
inspect it for damage. Immediately report any damage to the shipping agent.  
3.2 Mounting the Control Unit  
The Model 3010MB Control Unit is for indoor/outdoor use in a general  
purpose area. This Unit is NOT for any type of hazardous environments.  
The standard model is designed for indoor/outdoor mounting. Figure 3-  
1 is an illustration of a Model 3010MB standard Control Unit front panel  
and mounting brackets located-two at the top and two at the bottom of the  
unitsframe.  
Part I: 3-1  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3 Installation  
Model 3010MB  
NPT Fitting  
supplied by  
customer  
0.0  
AL-1  
%
AC POWER IN  
50/60 HZ  
3/4" NPT  
3/4" NPT  
100-240V  
ALARM OUTPUTS  
DIGITAL INPUT SPAN ZERO  
CAL. CONTACT RANGE  
ID CONTACTS RS-232  
1" NPT  
1" NPT  
SOLENOID RETUR  
ANALOG OUTPUTS  
REMOTE SENSOR  
NET WORK  
Figure 3-1: Front Panel of the Model 3010MB Control Unit  
All operator controls are mounted on the inner control panel, which is  
hinged on the left edge and doubles as a door to provide access to the inter-  
nal components of the instrument. The door will swing open when the  
button of the latch is pressed all the way in with a narrow gauge tool (less  
than 0.18 inch wide), such as a small hex wrench or screwdriver Allow  
3-2: Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Paramagnetic Oxygen Analyzer  
Part I: Control Unit  
clearance for the door to open in a 90-degree arc of radius 11.75 inches. See  
Figure 3-2.  
Figure 3-2: Required Front Door Clearance  
3.3 Electrical Connections  
Figure 3-3 shows the Control Unit interface panel. Connections for  
power, communications, and both digital and analog signal outputs are  
described in the following paragraphs. Wire size and maximum length data  
appear in the Drawings at the back of this manual.  
Figure 3-3: Interface Panel of the Model 3010MB Control Unit  
For safe connections, ensure that no uninsulated wire extends outside of  
the terminal blocks that are attached to. Stripped wire ends must insert  
completely into terminal blocks. No uninsulated wiring should be able to  
come in contact with fingers, tools or clothing during normal operation.  
Part I: 3-3  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3 Installation  
Model 3010MB  
Primary Input Power: The universal power supply requires a 100-  
240V ac, 50/60 Hz power source. See Figure 3-4 for detailed connections.  
DANGER: Power is applied to the instrument's circuitry as  
long as the instrument is connected to the power  
source. The standby function switches power on or  
off to the displays and outputs only.  
H ot  
G round  
N eutral  
Turn cw to hold  
ccw to  
loosen w ire.  
Insert w ire  
here.  
Figure 3-4: Primary Input Power Connections  
Fuse Installation: The fuse holders accept 5 x 20 mm, 1.0 A, T  
type (slow blow) fuses. Fuses are not installed at the factory. Be sure to  
install the proper fuse as part of installation (See Fuse Replacement in  
chapter 5, maintenance.)  
Analog Outputs: There are eight DC output signal connectors on  
the ANALOG OUTPUTS terminal block. There are two connectors per  
output with the polarity noted. See Figure 3-5.  
The outputs are:  
0–1 V dc % of Range: Voltage rises linearly with increasing oxygen, from  
0 V at 0% to 1 V at 100%. (Full scale = 100%  
programmedrange.)  
0–1 V dc Range ID:  
0.25 V = Low Range, 0.5 V = Medium Range,  
0.75 V = High Range, 1 V = Air Cal Range.  
3-4: Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Paramagnetic Oxygen Analyzer  
Part I: Control Unit  
4–20 mA dc % Range: (-M Option) Current increases linearly with increas-  
ing oxygen, from 4 mA at 0% to 20 mA at full scale  
100%. (Full scale = 100% of programmed range.)  
4–20 mA dc Range ID: (-M Option) 8 mA = Low Range, 12 mA = Me-  
dium Range, 16 mA = High Range, 20 mA = Air  
Cal Range.  
Figure 3-5: Analog Output Connections  
Examples:  
The analog output signal has a voltage which depends on the oxygen  
concentration AND the currently activated analysis range. To relate the  
signal output to the actual concentration, it is necessary to know what range  
the instrument is currently on, especially when the analyzer is in the  
autorangingmode.  
The signaloutput for concentration is linear over currently selected  
analysis range. For example, if the analyzer is set on a range that was  
defined as )-10 % O2, then the output would be as shown in Table 3-1.  
Part I: 3-5  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3 Installation  
Model 3010MB  
Table 3-1: Analog Concentration Output-Examples  
Voltage Signal  
Output (V dc)  
Current Signal  
Output (mA dc)  
%O2  
0
0.0  
0.1  
0.2  
0.3  
0.4  
0.5  
0.6  
0.7  
0.8  
0.9  
1.0  
4.0  
1
2
5.6  
7.2  
3
8.8  
4
10.4  
12.0  
13.6  
15.2  
16.8  
18.4  
20.0  
5
6
7
8
9
10  
To provide an indication of the range, a second pair of analog output  
terminals are used. They generate a steady preset voltage (or current when  
using the current outputs) to represent a particular range. Table 3-2 gives the  
range ID output for each analysis range.  
Table 3-2: Analog Range ID Output - Example  
Range  
LO  
Voltage (V)  
0.25  
Current (mA)  
8
MED  
HI  
0.50  
12  
16  
20  
0.75  
CAL (0-25%)  
1.00  
3-6: Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Paramagnetic Oxygen Analyzer  
Part I: Control Unit  
Alarm Relays:  
There are three alarm-circuit connectors on the alarm relays block  
(under RELAY OUTPUTS) for making connections to internal alarm relay  
contacts. Each provides a set of Form C contacts for each type of alarm.  
Each has both normally open and normally closed contact connections. The  
contact connections are indicated by diagrams on the rear panel. They are  
capable of switching up to 3 ampers at 250 V AC into a resistive load  
(Figure3-6).  
Figure 3-5: Types of Relay Contacts  
The connectors are:  
Threshold Alarm 1: • Can be configured as high (actuates when  
concentration is above threshold), or low  
(actuates when concentration is below thresh old).  
• Can be configured as fail-safe or non-fail-safe.  
• Can be configured as latching or nonlatching.  
• Can be configured out (defeated).  
Threshold Alarm 2: • Can be configured as high (actuates when concen-  
tration is above threshold), or low (actuates when  
concentrationisbelowthreshold).  
• Can be configured as fail-safe or non-fail-safe.  
• Can be configured as latching or nonlatching.  
• Can be configured out (defeated).  
SystemAlarm:  
Actuates when DC power supplied to circuits is  
unacceptable in one or more parameters. Permanently  
configured as fail-safe and latching. Cannot be de-  
feated. Actuates if self test fails.  
To reset a System Alarm during installation, discon-  
nect power to the instrument and then reconnect it  
Further detail can be found in chapter 4, section 4-5.  
Part I: 3-7  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3 Installation  
Model 3010MB  
Digital Remote Cal Inputs  
Remote Zero and Span Inputs: The REMOTE SPAN and RE-  
MOTE ZERO inputs are on the DIGITAL INPUT terminal block. They  
accept 0 V (OFF) or 24 V dc (ON) for remote control of calibration (See  
Remote Calibration Protocol below.)  
Zero:  
Floating input. 5 to 24 V input across the + and – terminals  
puts the analyzer into the ZERO mode. Either side may be  
grounded at the source of the signal. 0 to 1 volt across the  
terminals allows ZERO mode to terminate when done. A  
synchronous signal must open and close the external zero  
valve appropriately. See Remote Probe Connector at end of  
section 3.3. (With the -C option, the internal valves automati-  
callyoperatesynchronously).  
Span:  
Floating input. 5 to 24 V input across the + and – terminals  
puts the analyzer into the SPAN mode. Either side may be  
grounded at the source of the signal. 0 to 1 volt across the  
terminals allows SPAN mode to terminate when done. A  
synchronous signal must open and close the external span  
valve appropriately. See Remote Probe Connector at end of  
section 3.3. (With the -C option, the internal valves automati-  
callyoperatesynchronously.)  
Cal Contact: This relay contact is closed while analyzer is spanning  
and/or zeroing. (See Remote Calibration Protocol below.)  
Remote Calibration Protocol: To properly time the Digital Remote  
Cal Inputs to the Model 3010MB Analyzer, the customer's controller must  
monitor the Cal Relay Contact.  
When the contact is OPEN, the analyzer is analyzing, the Remote Cal  
Inputs are being polled, and a zero or span command can be sent.  
When the contact is CLOSED, the analyzer is already calibrating. It  
will ignore your request to calibrate, and it will not remember that request.  
Once a zero or span command is sent, and acknowledged (contact  
closes), release it. If the command is continued until after the zero or span is  
complete, the calibration will repeat and the Cal Relay Contact (CRC) will  
closeagain.  
Forexample:  
1) Test the CRC. When the CRC is open, Send a zero command  
until the CRC closes (The CRC will quickly close.)  
2) When the CRC closes, remove the zero command.  
3) When CRC opens again, send a span command until the CRC  
closes. (The CRC will quickly close.)  
3-8: Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Paramagnetic Oxygen Analyzer  
Part I: Control Unit  
4) When the CRC closes, remove the span command.  
When CRC opens again, zero and span are done, and the sample is  
being analyzed.  
Note: The Remote Probe connector (paragraph 3.3) provides signals  
to ensure that the zero and span gas valves will be controlled  
synchronously. If you have the -C Internal valve option -  
which includes additional zero and span gas inputs - the  
3010MB automatically regulates the zero, span and sample gas  
flow.  
Range ID Relays: Four dedicated RANGE ID CONTACT relays .  
The first three ranges are assigned to relays in ascending order—Low range  
is assigned to RANGE 1 ID, Medium range is assigned to RANGE 2 ID,  
and High range is assigned to RANGE 3 ID. RANGE 4 ID is reserved for  
the Air Cal Range (25%).  
Network I/O: A serial digital input/output for local network protocol.  
At this printing, this port is not yet functional. It is to be used in future  
versionsof the instrument.  
RS-232 Port: The digital signal output is a standard RS-232 serial  
communications port used to connect the analyzer to a computer, terminal, or  
other digital device. The pinouts are listed in Table 3-3.  
Table 3-3: RS-232 Signals  
RS-232 Sig RS-232 Pin Purpose  
DCD  
RD  
1
2
3
4
5
6
7
8
9
Data Carrier Detect  
Received Data  
TransmittedData  
Data Terminal Ready  
Common  
TD  
DTR  
COM  
DSR  
RTS  
CTS  
RI  
Data Set Ready  
Request to Send  
Clear to Send  
RingIndicator  
The data sent is status information, in digital form, updated every two  
seconds. Status is reported in the following order:  
Part I: 3-9  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3 Installation  
Model 3010MB  
The concentration in percent  
The range is use (HI< MED< LO)  
The span of the range 0-100%, etc)  
Which alarm - if any - are disabled (AL-x DISABLED)  
Which alarms - if any - are tripped (AL-x ON)  
Each status output is followed by a carriage return and line feed.  
Three input functions using RS-232 have been implemented to date.  
They are described in Table 3-4.  
Table 3-4: Commands via RS-232 Input  
Command  
as<enter>  
az<enter>  
st<enter>  
Description  
Immediatelystartsanautospan.  
Immediatelystartsanautozero.  
Toggling input. Stops/Starts any status message output  
from the RS-232, Until st<enter> is sent again.  
The RS-232 protocol allows some flexibility in its implementation.  
Table 3-5 lists certain RS-232 values that are required by the 3010MB.  
Table 3-5: Required RS-232 Options  
Parameter  
Baud  
Setting  
2400  
Byte  
Parity  
8 bits  
none  
Stop Bits  
MessageInterval  
1
2 seconds  
Remote Sensor and Solenoid Valves: The 3010MB is a single-  
chassis instrument. However, the REMOTE SENSOR and SOLENOID  
RETURN connectors are provided for use with a remote sensor and/or  
sampling system, if desired. See Figure 3-7 and 3-8.  
3-10: Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Paramagnetic Oxygen Analyzer  
Part I: Control Unit  
Thermistor 1  
Thermistor  
Sensor Signal 8  
Thermistor 2  
Block  
Sensor Signal 2  
Sensor Return (-)  
Sensor  
Sensor Signal 1  
Sensor Signal 7  
Block  
Sensor Hot (+)  
Figure 3-7 Remote Sensor Connector Pinouts  
+15 V dc  
+
-
Solenoid Valve 3  
Solenoid Valve 4  
Sample In  
Sample (return)  
+15 V dc  
+
-
Solenoid Valve 9  
Solenoid Valve 10  
Span In  
Zero In  
Exhaust  
Span (return)  
CU  
AU  
+15 V dc  
+
-
Solenoid Valve 5  
Solenoid Valve 6  
Zero (return)  
+15 V dc  
+
-
Solenoid Valve n/a  
Solenoid Valve n/a  
Exhaust (return)  
Figure 3-8: Remote Solenoid Return Connector Pinouts  
The voltage from the solenoid outputs is nominally 0 V for the OFF  
and 15 V dc for the ON conditions. The maximum combined current that  
can be pulled from these output lines is 100 mA. (If two lines are ON at the  
same time, each must be limited to 50 mA, etc.) If more current and/or a  
different voltage is required, use relays, power amplifiers, or other matching  
circuitry to provide the actual driving current.  
Note that each individual line has a series FET with a nominal ON  
resistance of 5 ohms (9 ohms worst case). This can limit the obtainable  
voltage, depending on the load impedance applied. See Figure 3-9.  
Figure 3-9: FET Series Resistance  
Part I: 3-11  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3 Installation  
Model 3010MB  
3.4 Testing the System  
After The Control Unit and the Analysis Unit are both installed and  
interconnected, and the system gas and electrical connections are complete,  
the system is ready to test. Before plugging either of the units into their  
respectivepowersources:  
Check the integrity and accuracy of the gas connections. Make  
sure there are no leaks.  
Check the integrity and accuracy of all electrical connections.  
Make sure there are no exposed conductors  
Check that sample pressure is between 3 and 40 psig, according  
to the requirements of your process.  
Power up the system, and test it by performing the following  
operations:  
1. Repeat the Self-Diagnostic Test as described in chapter 4, section  
4.3.5.  
3-12: Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Paramagnetic Oxygen Analyzer  
Part I: Control Unit  
Operation  
4.1 Introduction  
Once the analyzer has been installed, configure it for your process. To  
do this you can:  
Setsystemparameters—  
Specify a password, if desired, requiring operator to log in.  
Establish and start an automatic calibration cycle, if desired.  
Calibratetheinstrument.  
Define the three user selectable analysis ranges. Then choose  
autoranging or select a fixed range of analysis, as required.  
Set alarm setpoints, and modes (latching, failsafe, etc).  
Before configuration these default values are in effect:  
PARAMETER  
DEFAULT  
LO Range  
100 %  
MED Range  
HI Range  
1000 %  
10,000 %  
Auto Ranging  
Alarm Relays  
ON  
1000 %  
(Defeated, HI, Not failsafe, Not latching)  
000008.00 %  
Span  
(Auto, every 0 days at 0 hours)  
(Auto, every 0 days at 0 hours).  
Zero  
If you choose not to use password protection, the default password is  
automatically displayed on the password screen when you start up, and you  
simply press Enter for access to all functions of the analyzer.  
Part I: 4-1  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4 Operation  
Model 3010MB  
4.2 Using the Data Entry and Function  
Buttons  
Data Entry Buttons: The < > arrow buttons select options from the  
menu currently being displayed on the VFD screen. The selected option  
blinks.  
When the selected option includes a modifiable item, the arrow  
buttons can be used to increment or decrement that modifiable item.  
The Enter button is used to accept any new entries on the VFD screen.  
The Escape button is used to abort any new entries on the VFD screen that  
are not yet accepted by use of the Enter button.  
Figure 4-1 shows the hierarchy of functions available to the operator via  
the function buttons. The six function buttons on the analyzer are:  
Analyze. This is the normal operating mode. The analyzer  
monitors the oxygen content of the sample, displays the  
concentration of oxygen, and warns of any alarm conditions.  
System. The system function consists of six subfunctions that  
regulate the internal operations of the analyzer:  
Auto-Cal setup  
Passwordassignment  
Self-Testinitiation  
Checking software version  
Logging out.  
Zero. Usedto set upa zero calibration.  
Span. Used to set up a span calibration.  
Alarms. Used to set the alarm setpoints and determine whether  
each alarm will be active or defeated, HI or LO acting, latching,  
and/orfailsafe.  
Range. Used to set up three analysis ranges that can be switched  
automatically with autoranging or used as individual fixed  
ranges.  
Any function can be selected at any time by pressing the appropriate  
button (unless password restrictions apply). The order as presented in this  
manual is appropriate for an initial setup.  
4-2: Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Paramagnetic Oxygen Analyzer  
Part I: Control Unit  
ANALYZE  
SYSTE  
Perform Oxygen  
Analysis of  
the Sample  
SPAN  
TRAK/HLD  
ZERO  
Set Instrument  
Span  
ALARMS  
Perform  
Self-Diagnostic  
Test  
Set Instrument  
Zero  
Initiate  
Automatic  
Calibratio  
Set Alarm  
Setpoints  
Confrigure Mode  
of Alarm  
Operation  
Set Password  
Logout  
Show Negative  
Figure 4-1: Hierarchy of Functions and Subfunctions  
Each of these functions is described in greater detail in the following  
procedures. The VFD screen text that accompanies each operation is repro-  
duced, at the appropriate point in the procedure, in a Monospaced type  
style. Pushbutton names are printed in Oblique type.  
4.3 The System Function  
The subfunctions of the System function are described below. Specific  
procedures for their use follow the descriptions:  
Auto-Cal: Used to define an automatic calibration sequence  
and/or start an Auto-Cal.  
PSWD: Security can be established by choosing a 5 digit  
password (PSWD) from the standard ASCII character set. (See  
Installing or Changing a Password, below, for a table of ASCII  
characters available.) Once a unique password is assigned and  
Part I: 4-3  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4 Operation  
Model 3010MB  
activated, the operator MUST enter the UNIQUE password to  
gain access to set-up functions which alter the instrument's  
operation, such as setting the instrument span or zero setting,  
adjusting the alarm setpoints, or defining analysis ranges.  
After a password is assigned, the operator must log out to  
activate it. Until then, anyone can continue to operate the  
instrument without entering the new password.  
Only one password can be defined. Before a unique password  
is assigned, the system assigns TETAI by default. This allows  
access to anyone. After a unique password is assigned, to defeat  
the security, the password must be changed back to TETAI.  
Logout: Logging out prevents an unauthorized tampering with  
analyzersettings.  
More: Select and enter More to get a new screen with additional  
subfunctionslisted.  
Self–Test: The instrument performs a self-diagnostic test to  
check the integrity of the power supply, output boards and  
amplifiers.  
Version: Displays Manufacturer, Model, and Software Version  
ofinstrument.  
Showing Negative: The operator selects whether display can  
show negative readings or not.  
TRAK/HLD: The operator sets whether the instrument analog  
outputs track the concentration change during calibration and sets  
a time delay for the concentration alarms after calibration  
4.3.1 Tracking the Oxygen Readings during Calibration  
and Alarm delay  
The user has the option of setting the preferenc as to whether the analog  
outputs track the display readings during calibration or not. To set the prefer-  
ence, press the System key once and the first System menu will appear in the  
VFD display:  
TRAK/HLD Auto-Cal  
PSWD Logout More  
TRAK/HLD should be blinking. To enter this system menu press the  
Enter key once:  
Output Sttng: TRACK  
Alarm Dly: 10 min  
4-4: Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Paramagnetic Oxygen Analyzer  
Part I: Control Unit  
Or  
Output Sttng: HOLD  
Alarm Dly: 10 min  
In the first line, TRACK or HOLD should be blinking. The operator  
can toggle between TRACK and HOLD with the Up or Down keys. When  
TRACK is selected, the analog outputs (0-1 VDC and 4-20 ma) and the  
range ID contacts will track the instrument readings during calibration (either  
zero or span). TRACK is the factory default.  
When HOLD is selected, the analog outputs (0-1 VDC and 4-20 ma)  
and the range ID contacts will freeze on their last state before entering one of  
the calibration modes. When the instrument returns to the Analyze mode,  
either by a successful or an aborted calibration, there will be a three-minute  
delay before the analog outputs and the range ID contacts start tracking  
again.  
The concentration alarms freeze on their last state before entering  
calibration regardless of selecting HOLD or TRACK. But, when HOLD is  
selected the concentration alarms will remain frozen for the time displayed in  
the second line of the TRAK/HLD menu after the analyzer returns to the  
Analyze mode.  
The factory default is three minutes, but the delay time is program-  
mable. To adjust to delay time use the Left or Right arrow keys. When the  
time displayed on the second line blinks, it can be adjusted by Pressing the  
Up or Down keys to increase or decrease its value. The minimum delay is 1  
minute, the maximum is 30.  
This preference is stored in non-volatile memory so that it is recovered  
if power is removed from the instrument.  
4.3.2 Setting up an Auto-Cal  
When the proper calibration gases are connected (see chapter 3, instal-  
lation), the Analyzer can cycle itself through a sequence of steps that auto-  
matically zero and span the instrument.  
Note: If you require highly accurate Auto-Cal timing, use external  
Auto-Cal control where possible. The internal clock in the  
Model 3010MB is accurate to 2-3 %. Accordingly, internally  
scheduled calibrations can vary 2-3 % per day.  
To setup an Auto–Cal cycle:  
Choose System from the Function buttons. The VFD will display five  
subfunctions.  
Part I: 4-5  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4 Operation  
Model 3010MB  
TRAK/HLD Auto—Cal  
PSWD Logout More  
Use < > arrows to blink Auto—Cal, and press Enter. A new screen for  
Span/Zero set appears.  
SpanOFFNxt: 0d0h  
ZeroOFFNxt: 0d0h  
Press < > arrows to blink Span (or Zero), then press Enter again. (You  
won’t be able to set OFF to ON if a zero interval is entered.) A Span  
Every ... (or Zero Every ...) screen appears.  
SpanEvery0 d  
Start0hfromnow  
Use arrows to set an interval value, then use < > arrows to move to  
the start-time value. Use arrows to set a start-time value.  
To turn ON the Span and/or Zero cycles (to activate Auto-Cal): Press  
System again, choose Auto—Cal, and press Enter again. When the Span/  
Zero values screen appears, use the < > arrows to blink the Span (or Zero)  
OFF/ON field. Use arrows to set the OFF/ON field to ON. You can  
now turn these fields ON because there is a nonzero span interval defined.  
4.3.3 Password Protection  
If a password is assigned, then setting the following system parameters  
can be done only after the password is entered: span and zero settings,  
alarm setpoints, analysis range definitions, switching between autoranging  
and manual override, setting up an auto-cal, and assigning a new password.  
However, the instrument can still be used for analysis or for initiating a self-  
test without entering the password.  
If you have decided not to employ password security, use the default  
password TETAI. This password will be displayed automatically by the  
microprocessor. The operator just presses the Enter key to be allowed total  
accesstotheinstrument’sfeatures.  
NOTE: If you use password security, it is advisable to keep a copy of  
the password in a separate, safe location.  
4.3.3.1  
EnteringthePassword  
To install a new password or change a previously installed password,  
you must key in and ENTER the old password first. If the default password  
4-6: Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Paramagnetic Oxygen Analyzer  
Part I: Control Unit  
is in effect, pressing the ENTER button will enter the default TETAI pass-  
word for you.  
Press System to enter the System mode.  
TRAK/HLD Auto—Cal  
PSWD Logout More  
Use the < > arrow keys to scroll the blinking over to PSWD, and press  
Enter to select the password function. Either the default TBEAI password or  
AAAAA place holders for an existing password will appear on screen  
depending on whether or not a password has been previously installed.  
TET AI  
Enter PWD  
or  
AAAAA  
Enter PWD  
The screen prompts you to enter the current password. If you are not  
using password protection, press Enter to accept TETAI as the default  
password. If a password has been previously installed, enter the password  
using the < > arrow keys to scroll back and forth between letters, and the ∇  
arrow keys to change the letters to the proper password. Press Enter to enter  
the password.  
If the password is accepted, the screen will indicate that the password  
restrictions have been removed and you have clearance to proceed.  
PSWDRestrictions  
Removed  
In a few seconds, you will be given the opportunity to change this  
password or keep it and go on.  
ChangePassword?  
<ENT>=Yes <ESC>=No  
Press Escape to move on, or proceed as in Changing the Password,  
below.  
4.3.3.2  
InstallingorChangingthePassword  
If you want to install a password, or change an existing password,  
proceed as above in Entering the Password. When you are given the oppor-  
tunity to change the password:  
ChangePassword?  
<ENT>=Yes <ESC>=No  
Part I: 4-7  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4 Operation  
Model 3010MB  
Press Enter to change the password (either the default TETAI or the  
previously assigned password), or press Escape to keep the existing pass-  
word and move on.  
If you chose Enter to change the password, the password assignment  
screenappears.  
TET AI  
<ENT>ToProceed  
or  
AAAAA  
<ENT>ToProceed  
Enter the password using the < > arrow keys to move back and forth  
between the existing password letters, and the arrow keys to change the  
letters to the new password. The full set of 94 characters available for pass-  
word use are shown in the table below.  
Characters Available for Password Definition:  
A
K
U
_
i
s
}
)
3
=
B
L
V
`
j
t
*
4
>
C
M
W
a
k
u
!
+
5
?
D
N
X
b
l
v
"
'
6
@
E
O
Y
c
m
w
#
-
7
F
P
Z
d
n
x
$
.
8
G
Q
[
e
o
y
%
/
9
H
R
¥
f
p
z
&
0
:
I
S
]
g
q
{
'
1
;
J
T
^
h
r
|
(
2
<
When you have finished typing the new password, press Enter. A  
verification screen appears. The screen will prompt you to retype your  
passwordforverification.  
AAAAA  
RetypePWDToVerify  
Wait a moment. The entry screen will give you clearance to proceed.  
AAAAA  
<ENT> TO Proceed  
Use the arrow keys to retype your password and press Enter when  
finished. Your password will be stored in the microprocessor and the system  
will immediately switch to the Analyze screen, and you now have access to  
allinstrumentfunctions.  
4-8: Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Paramagnetic Oxygen Analyzer  
Part I: Control Unit  
If no alarms are tripped, the Analyze screen appears as:  
0.0  
ppm AnlZ  
Range: 0100  
If an alarm is tripped, the second line will change to show which alarm  
itis:  
0.0 ppm Anlz  
AL—1  
NOTE:If you previously logged off the system , you will now be  
required to re-enter the password to gain access to Span,  
Zero, Alarm, and Range functions.  
4.3.4 Logout  
The Logout function provides a convenient means of leaving the  
analyzer in a password protected mode without having to shut the instrument  
off. By entering Logout, you effectively log off the instrument leaving the  
system protected against use until the password is reentered. To log out,  
press the System button to enter the System function.  
TRAK/HLD Auto—Cal  
PSWD Logout More  
Use the < > arrow keys to position the blinking over the Logout func-  
tion, and press Enter to Log out. The screen will display the message:  
ProtectedUntil  
PasswordReentered  
4.3.5 System Self-Diagnostic Test  
The Model 3010MB has a built-in self-diagnostic testing routine. Pre-  
programmed signals are sent through the power supply, output board and  
sensor circuit. The return signal is analyzed, and at the end of the test the  
status of each function is displayed on the screen, either as OK or as a  
number between 1 and 3. (See System Self Diagnostic Test in chapter 5 for  
number code.)  
Note: Remote Probe connector must be connected to the Analysis  
Unit, or sensor circuit will not be properly checked.  
The self diagnostics are run automatically by the analyzer whenever the  
instrument is turned on, but the test can also be run by the operator at will.  
To initiate a self diagnostic test during operation:  
Part I: 4-9  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4 Operation  
Model 3010MB  
Press the System button to start the System function.  
TRAK/HLD Auto—Cal  
PSWD Logout More  
Use the < > arrow keys to blink More, then press Enter.  
VersionSelf—Test  
Use the < > arrow keys again to move the blinking to the Self–Test  
function. The screen will follow the running of the diagnostic.  
RUNNING DIAGNOSTIC  
TestingPreamp83  
During preamp testing there is a countdown in the lower right corner of  
the screen. When the testing is complete, the results are displayed.  
Power:OK Analog:OK  
Preamp:3  
The module is functioning properly if it is followed by OK. A number  
indicates a problem in a specific area of the instrument. Refer to chapter 5  
Maintenance for number-code information. The results screen alternates for a  
timewith:  
PressAnyKey  
ToContinue...  
Then the analyzer returns to the initial System screen.  
4.3.6 Version Screen  
Move the < > arrow key to More and press Enter. With Version  
blinking, press Enter. The screen displays the manufacturer, model, and  
softwareversioninformation.  
4.3.7 Showing Negative Oxygen Readings  
For software version 1.4.4 or later, the instrument only displays oxygen  
readings that are positive or zero. The instrument can be reconfigured to  
show negative readings if sensor output drifts below zero. This situation  
may arise after the instrument has been zeroed, as time progresses the sensor  
may drift below the zero calibration setpoint.  
To show negative oxygen readings on the display:  
- Press the System key  
4-10: Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Paramagnetic Oxygen Analyzer  
Part I: Control Unit  
TRAK/HLD Auto-Cal  
PSWD Logout More  
- Use the Right or Left arrow keys and select More. Press Enter.  
Version Self-Test  
Show_Negative=NO  
- Use the Right or Left arrow keys and select  
“Show_Negative=NO”.  
- Use the Up or Down key to toggle from NO to YES.  
- Press the Escape key twice to return to the analyze mode.  
This preference is stored in non-volatile memory, so this configuration  
is remembered after a power shutdown. If the instrument is cold started, it  
will go back to default (not showingg negative oxygen readings).  
4.4 The Zero and Span Functions  
Zeroing is not required in order to achieve the published  
accuracy specification of this unit.  
Zeroing will eliminate offset error contributed by sensor,  
electronics, andinternalandexternalsamplingsystemand  
improve performance beyond published specification lim-  
its.  
The analyzer is calibrated using zero and span gases.  
Any suitable oxygen-free gas can be used for zero gas as long as it is  
known that it will not react adversely with the sample system.  
Although the instrument can be spanned using air, a span gas with a  
known oxygen concentration in the range of 70–90% of full scale of the  
range of interest is recommended. Since the oxygen concentration in air is  
20.9 % (209,000 ppm), the cell can take longer to recover if the instrument is  
used for trace or less than 1% full scale oxygen analysis immediately  
followingcalibrationinair.  
Connect the calibration gases to the analyzer according to the instruc-  
tions given in Section 3.4.1, Gas Connections, observing all the prescribed  
precautions.  
Shut off the gas pressure before connecting it to the analyzer, and  
be sure to limit the pressure to 40 psig or less when turning it back on.  
Readjust the gas pressure into the analyzer until the flowrate (as read on  
the Analysis Unit SLPM flowmeter) settles between 0.5 and 2.4 SLPM  
(approximately1-5scfh).  
Part I: 4-11  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4 Operation  
Model 3010MB  
If you are using password protection, you will need to enter your  
password to gain access to either of these functions. Follow the instructions  
in sections 4.3.3.2 or 4.3.3.3 to enter your password. Once you have gained  
clearance to proceed, you can enter the Zero or Span function.  
4.4.1 Zero Cal  
The Zero button on the front panel is used to enter the zero calibration  
function. Zero calibration can be performed in either the automatic or manual  
mode. In the automatic mode, an internal algorithm compares consecutive  
readings from the sensor to determine when the output is within the accept-  
able range for zero. In the manual mode, the operator determines when the  
reading is within the acceptable range for zero. Make sure the zero gas is  
connected to the instrument. If you get a CELL FAILURE message skip to  
section 4.4.1.3.  
4.4.1.1  
AutoModeZeroing  
Press Zero to enter the zero function mode. The screen allows you to  
select whether the zero calibration is to be performed automatically or manu-  
ally. Use the arrow keys to toggle between AUTO and MAN zero  
settling. Stop when AUTO appears, blinking, on the display.  
Zero:Settling:AUTO  
<ENT>ToBegin  
Press Enter to begin zeroing.  
#### PPM Zero  
Slope=#### ppm/s  
The beginning zero level is shown in the upper left corner of the dis-  
play. As the zero reading settles, the screen displays and updates information  
on Slope (unless the Slope starts within the acceptable zero range and does  
not need to settle further).  
Then, and whenever Slope is less than 0.08 for at least 3 minutes,  
instead of Slope you will see a countdown: 5 Left, 4 Left, and so fourth.  
These are five steps in the zeroing process that the system must complete,  
AFTER settling, before it can go back to Analyze.  
#### PPM Zero  
4Left=### ppm/s  
The zeroing process will automatically conclude when the output is  
within the acceptable range for a good zero. Then the analyzer automatically  
returns to the Analyze mode.  
4-12: Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Paramagnetic Oxygen Analyzer  
Part I: Control Unit  
4.4.1.2  
ManualModeZeroing  
Press Zero to enter the Zero function. The screen that appears allows  
you to select between automatic or manual zero calibration. Use the keys  
to toggle between AUTO and MAN zero settling. Stop when MAN appears,  
blinking, on the display.  
Zero: Settling: Man  
<ENT>ToBegin  
Press Enter to begin the zero calibration. After a few seconds the first  
of five zeroing screens appears. The number in the upper left hand corner is  
the first-stage zero offset. The microprocessor samples the output at a prede-  
termined rate. It calculates the differences between successive samplings and  
displays the rate of change as Slope= a value in parts per million per second  
(ppm/s).  
#### ppm Zero  
Slope=#### ppm/s  
NOTE:It takes several seconds for the true Slope value to display.  
Wait about 10 seconds. Then, wait until Slope is sufficiently  
close to zero before pressing Enter to finish zeroing. Slope is  
given in ppm/s.  
Generally, you have a good zero when Slope is less than 0.05 ppm/s  
for about 30 seconds. When Slope is close enough to zero, press Enter. In a  
few seconds, the screen will update.  
Once span settling completes, the information is stored in the  
microprocessor, and the instrument automatically returns to theAnalyze  
mode.  
4.4.1.3  
CellFailure  
Cell failure in the 3010MB is usually associated with inability to zero  
the instrument down to a satisfactorily low ppm reading, e.g. cell does not  
fail if it comes below 5PPM . When this occurs, the 3010MB system alarm  
trips, and the LCD displays a failure message.  
#.# ppm Anlz  
CELL FAIL/ ZERO HIGH  
Before replacing the cell:  
a. Check your span gas to make sure it is within specifications.  
b. Check for leaks downstream from the cell, where oxygen may be  
leaking into the system.  
Part I: 4-13  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4 Operation  
Model 3010MB  
c. Check whether more time is needed for readings to drop to a  
satisfactory level. This might happen when zero was started from  
very high PPM level  
If there are no leaks and the span gas is OK, replace the cell as de-  
scribed in Part II Analysis Units, chapter 5 Maintenance.  
4.4.2 Span Cal  
The Span button on the front panel is used to span calibrate the ana-  
lyzer. Span calibration can be performed using the automatic mode, where  
an internal algorithm compares consecutive readings from the sensor to  
determine when the output matches the span gas concentration. Span cali-  
bration can also be performed in manual mode, where the operator deter-  
mines when the span concentration reading is acceptable and manually exits  
thefunction.  
4.4.2.1  
AutoModeSpanning  
Press Span to enter the span function. The screen that appears allows  
you to select whether the span calibration is to be performed automatically or  
manually. Use the arrow keys to toggle between AUTO and MAN span  
settling. Stop when AUTO appears, blinking, on the display.  
Span:Settling:AUTO  
<ENT>ForNext  
Press Enter to move to the next screen.  
Calib.Holdingtime  
Calhold: 5 min  
This menue allows the operator to set the time the analyzer should be  
held in the auto span mode. It does not affect anything in Manual Mode.  
Just press Enter to continue.  
Press Enter to move to the next screen.  
SpanVal:000008.00  
<ENT>Span <UP>Mod #  
Use the arrow keys to enter the oxygen-concentration mode. Use  
the < > arrow keys to blink the digit you are going to modify. Use the ∇  
arrow keys again to change the value of the selected digit. When you have  
finished typing in the concentration of the span gas you are using  
(209000.00 if you are using air), press Enter to begin the Span calibration.  
4-14: Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Paramagnetic Oxygen Analyzer  
Part I: Control Unit  
####  
ppm  
Span  
Slope=#### ppm/s  
The beginning span value is shown in the upper left corner of the  
display. As the span reading settles, the screen displays and updates informa-  
tion on Slope. Spanning automatically ends when the span output corre-  
sponds, within tolerance, to the value of the span gas concentration. Then the  
instrumentautomaticallyreturnstotheanalyzemode.  
4.4.2.2  
ManualModeSpanning  
Press Span to start the Span function. The screen that appears allows  
you to select whether the span calibration is to be performed automatically or  
manually.  
Span:Settling:MAN  
<ENT>ForNext  
Use the keys to toggle between AUTO and MAN span settling.  
Stop when MAN appears, blinking, on the display. Press Enter to move to  
the next screen.  
Press Enter to move to the next screen.  
Calib.Holdingtime  
Calhold: 5 min  
This menue allows the operator to set the time the analyzer should be  
held in the auto span mode. It does not affect anything in the Manual Mode.  
Just press Enter to continue.  
SpanVal:000008.00  
<ENT>Span <UP>Mod #  
Press(<UP>)topermitmodification(Mod#)ofspanvalue.  
Use the arrow keys to enter the oxygen concentration of the span gas  
you are using (209000.00 if you are using air). The < > arrows choose the  
digit, and the arrows choose the value of the digit.  
Press Enter to enter the span value into the system and begin the span  
calibration.  
Once the span has begun, the microprocessor samples the output at a  
predeterminedrate. Itcalculatesthedifferencebetweensuccessivesamplings  
and displays this difference as Slope on the screen. It takes several seconds  
for the first Slope value to display. Slope indicates rate of change of the Span  
reading. It is a sensitive indicator of stability.  
Part I: 4-15  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4 Operation  
Model 3010MB  
####  
% Span  
Slope=#### ppm/s  
When the Span value displayed on the screen is sufficiently stable,  
press Enter. (Generally, when the Span reading changes by 1 % or less of  
the full scale of the range being calibrated for a period of ten minutes it is  
sufficiently stable.) Once Enter is pressed, the Span reading changes to the  
correct value. The instrument then automatically enters the Analyze func-  
tion.  
4.4.3 Span Failure  
The analyzer checks the output of the cell at the end of the span. If the  
raw output of the cell is less than 0.5 uA/ppm O2, the span will not be  
accepted. The analyzer will return to the previous calibration values, trigger  
the System Alarm, and display in the VFD:  
SpanFailed!!  
This message will be shown for five seconds and the instrument shall  
return to the Analyze mode. In the upper right hand corner of the VFD  
display “FCAL” will be shown. This message flag will help the operator  
troubleshoot in case calibration was initiated remotely. To reset the alarm  
and the flag message, the unit must be turned off by cycling the standby key  
. It will not reset if the next span cycle is correct.  
A trace cell is unlikely to fail span. As explained before, when the  
sensor reaches the end of its useful life, the zero offset begins to rise until the  
analyzer finds the zero unsatisfactory. Nevertheless, feeding the wrong span  
gas or electronics failure could set this feature off at the end of the span.  
Consider this before replacing the cell.  
4.5 The Alarms Function  
The Model 3010MB is equipped with 2 fully adjustable concentration  
alarms and a system failure alarm. Each alarm has a relay with a set of form  
C contacts rated for 3 amperes resistive load at 250 V ac. See figure in  
chapter 3, Installation and/or the Interconnection Diagram included at the  
back of this manual for relay terminal connections.  
The system failure alarm has a fixed configuration described in chapter  
3Installation.  
The concentration alarms can be configured from the front panel as  
either high or low alarms by the operator. The alarm modes can be set as  
4-16: Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Paramagnetic Oxygen Analyzer  
Part I: Control Unit  
latching or nonlatching, and either failsafe or nonfailsafe, or, they can be  
defeated altogether. The setpoints for the alarms are also established using  
thisfunction.  
Decide how your alarms should be configured. The choice will depend  
upon your process. Consider the following four points:  
1. Which if any of the alarms are to be high alarms and which if any  
are to be low alarms?  
Setting an alarm as HIGH triggers the alarm when the oxygen  
concentration rises above the setpoint. Setting an alarm as LOW  
triggers the alarm when the oxygen concentration falls below the  
setpoint.  
Decide whether you want the alarms to be set as:  
Both high (high and high-high) alarms, or  
One high and one low alarm, or  
Both low (low and low-low) alarms.  
2. Are either or both of the alarms to be configured as failsafe?  
In failsafe mode, the alarm relay de-energizes in an alarm  
condition. For nonfailsafe operation, the relay is energized in an  
alarm condition. You can set either or both of the concentration  
alarms to operate in failsafe or nonfailsafe mode.  
3. Are either of the alarms to be latching?  
In latching mode, once the alarm or alarms trigger, they will  
remain in the alarm mode even if process conditions revert back  
to no-alarm conditions. This mode requires an alarm to be  
recognized before it can be reset. In the nonlatching mode, the  
alarm status will terminate when process conditions revert to no-  
alarmconditions.  
4. Are either of the alarms to be defeated?  
The defeat alarm mode is incorporated into the alarm circuit so  
that maintenance can be performed under conditions which  
would normally activate the alarms.  
The defeat function can also be used to reset a latched alarm.  
(See procedures, below.)  
If you are using password protection, you will need to enter your  
password to access the alarm functions. Follow the instructions in Section  
4.3.3 to enter your password. Once you have clearance to proceed, enter the  
Alarm function.  
Press the Alarm button on the front panel to enter the Alarm function.  
Make sure that AL–1 is blinking.  
Part I: 4-17  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4 Operation  
Model 3010MB  
AL—1  
AL—2  
Choose Alarm  
Set up alarm 1 by moving the blinking over to AL–1 using the < >  
arrow keys. Then press Enter to move to the next screen.  
AL—11000 ppm HI  
Dft—N Fs—N Ltch—N  
Five parameters can be changed on this screen:  
Value of the alarm setpoint, AL–1 #### (ppm or % oxygen)  
Out-of-range direction, HI or LO  
Defeated? Dft–Y/N (Yes/No)  
Failsafe? Fs–Y/N (Yes/No)  
Latching? Ltch–Y/N (Yes/No).  
To define the setpoint, use the < > arrow keys to move the  
blinking over to AL–1 ####. Then use the arrow keys to  
change the number. Holding down the key speeds up the  
incrementingordecrementing. (Remember, setpointunitsare  
parts-per-million.)  
To set the other parameters use the < > arrow keys to move the  
blinking over to the desired parameter. Then use the arrow  
keys to change the parameter.  
Once the parameters for alarm 1 have been set, press Alarms  
again, and repeat this procedure for alarm 2 (AL–2).  
To reset a latched alarm, go to Dft– and then press either two  
times or two times. (Toggle it to Y and then back to N.)  
–OR –  
Go to Ltch– and then press either two times or two times.  
(Toggle it to N and back to Y.)  
4.6 The Range Function  
The Range function allows the operator to program up to three concen-  
tration ranges to correlate with the DC analog outputs. If no ranges are  
defined by the user, the instrument defaults to:  
Range  
Limits  
Low  
Med  
High  
0–1%  
0–5%  
0–10%  
4-18: Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Paramagnetic Oxygen Analyzer  
Part I: Control Unit  
The Model 3010MB is set at the factory to default to autoranging. In  
thismode, themicroprocessorautomaticallyrespondstoconcentration  
changes by switching ranges for optimum readout sensitivity. If the current  
range limits are exceeded, the instrument will automatically shift to the next  
higher range. If the concentration falls to below 85% of full scale of the next  
lower range, the instrument will switch to that range. A corresponding shift  
in the DC percent-of-range output, and in the range ID outputs, will be  
noticed.  
The autoranging feature can be overridden so that analog output stays  
on a fixed range regardless of the oxygen concentration detected. If the  
concentration exceeds the upper limit of the range, the DC output will  
saturate at 1 V dc (20 mA at the current output).  
However, the digital readout and the RS-232 output of the concentra-  
tion are unaffected by the fixed range. They continue to read accurately with  
full precision. See Front Panel description in chapter 1.  
The automatic air calibration range is always 0-25 % and is not pro-  
grammable.  
4.6.1 Setting the Analog Output Ranges  
To set the ranges, enter the range function mode by pressing the  
Range button on the front panel.  
L—### M—####  
H—##### Mode—AUTO  
Use the < > arrow keys to blink the range to be set: low (L), medium  
(M), or high (H).  
Use the arrow keys to enter the upper value of the range (all ranges  
begin at 0 ppm). Repeat for each range you want to set. Press Enter to  
accept the values and return to Analyze mode. (See note below.)  
Note: The ranges must be increasing from low to high, for example,  
if range 1 is set as 0–10% and range 2 is set as 0–100%, range  
3 cannot be set as 0–50% since it is lower than range 2.  
Part I: 4-19  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4 Operation  
Model 3010MB  
4.6.2 Fixed Range Analysis  
The autoranging mode of the instrument can be overridden, forcing the  
analyzer DC outputs to stay in a single predetermined range.  
To switch from autoranging to fixed range analysis, enter the range  
function by pressing the Range button on the front panel.  
Use the < > arrow keys to move the blinking over AUTO.  
Use the arrow keys to switch from AUTO to FX/LO, FX/MED, or  
FX/HI to set the instrument on the desired fixed range (low, medium, or  
high).  
L—### M—####  
H—##### Mode—FX/LO  
or  
L—### M—####  
H—##### Mode—FX/MED  
or  
L—### M—####  
H—##### Mode—FX/HI  
Press Escape to re-enter the Analyze mode using the fixed range.  
NOTE:When performing analysis on a fixed range, if the oxygen  
concentration rises above the upper limit (or default value) as  
established by the operator for that particular range, the  
output saturates at 1 V dc (or 20 mA). However, the digital  
readout and the RS-232 output continue to read the true value  
of the oxygen concentration regardless of the analog output  
range.  
4.7 The Analyze Function  
When the Analyze function is active, the 3010MB is monitoring the  
sample gas currently flowing in the Analysis Unit cell block. All undefeated  
alarms are ready to activate should their respective setpoints be crossed.  
Press the Analyze button to put the analyzer in the Analyze mode.  
Normally, all of the functions automatically switch back to the Analyze  
function when they have completed their assigned operations. Pressing the  
Escape button in many cases also switches the analyzer back to the Ana-  
lyze function. Alternatively, you can press the Analyze button at any time  
to return to analyzing your sample.  
4-20: Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Paramagnetic Oxygen Analyzer  
Part I: Control Unit  
4.8 Signal Output  
The standard Model 3010MB Percent Paramagnetic Oxygen Analyzer  
is equipped with two 0-1 V dc analog output terminals accessible on the  
interface panel (one concentration and one range ID) and two isolated 4-20  
mA dc current outputs (one concentration and one range ID).  
See Rear Panel in chapter 3, Installation, for illustration.  
The signal output for concentration is linear over the currently selected  
analysis range. For example, if the analyzer is set on range that was defined  
as 0–10% O2, then the output would be:  
Voltage Signal Current Signal  
ppm O2  
Output (V dc)  
Output (mA dc)  
0
1
2
3
4
5
6
7
8
0.0  
0.1  
0.2  
0.3  
0.4  
0.5  
0.6  
0.7  
0.8  
0.9  
1.0  
4.0  
5.6  
7.2  
8.8  
10.4  
12.0  
13.6  
15.2  
16.8  
18.4  
20.0  
9
10  
The analog output signal has a voltage which depends on the oxygen  
concentration AND the currently activated analysis range. To relate the  
signal output to the actual concentration, it is necessary to know what range  
the instrument is currently on, especially when the analyzer is in the  
autorangingmode.  
To provide an indication of the range, a second pair of analog output  
terminals are used. They generate a steady preset voltage (or current when  
using the current outputs) to represent a particular range. The following table  
gives the range ID output for each analysis range:  
Range  
Voltage (V)  
Current (mA)  
LO  
0.25  
8
MED  
0.50  
0.75  
1.00  
12  
16  
20  
HI  
CAL (0-25%)  
Part I: 4-21  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4 Operation  
Model 3010MB  
IMPORTANT:In the event of loss of flow through the analyzer, if the  
vent is vented to a location of high oxygen content,  
oxygen will back diffuse through the vent line and in  
most cases quickly saturate the cell with oxygen which  
can then require a quite long purge down time for  
the sensor when then exposed to low oxygen concen-  
trations. In the event that flow is to be interrupted into  
the analyzer, it is suggested that the user do one of the  
following:  
1.  
2.  
Bag the sensor in nitrogen during this time  
Install a shut off valve on the vent port of the ana-  
lyzer or somewhere within the users sample system.  
4-22: Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Part I: Control Unit  
Maintenance 5  
Maintenance  
Aside from normal cleaning and checking for leaks at the gas connec-  
tions, routine maintenance is limited to replacing fuses, and recalibration.  
Checking for leaks, and replacing fuses in the Analysis Unit are cov-  
ered in Part II, Chapter 5. For recalibration, see Part I, section 4.4 Calibra-  
tion.  
WARNING: SEE WARNINGS ON THE TITLE PAGE OF THIS  
MANUAL.  
5.1 Fuse Replacement  
The 3010MB requires two 5 x 20 mm, 1.0 A, T type (Slow Blow)  
fuses.  
The fuses are located inside the main housing on the Electrical  
Connector Panel, as shown in Figure 5-3. To replace a fuse:  
1. Disconnect the Unit from its power source.  
2. Place a small screwdriver in the notch in the fuse  
holder cap, push in, and rotate 1/4 turn. The cap will  
pop out a few millimeters. Pull out the fuse cap and  
fuse, as shown in Figure 5-1  
Part I: 5-1  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5 Maintenance  
Model 3010MB Percent Paramagnetic Oxygen Analyzer  
Figure 5-1: Removing Fuse Block Cap and Fuse from Housing  
2. Replace fuse by reversing process in step 1.  
5.2 System Self Diagnostic Test  
1. Press the System button to enter the system mode.  
2. Use the < > arrow keys to move to More, and press Enter.  
3. Use the < > arrow keys to move to Self-Test, and press Enter.  
The following failure codes apply:  
Table 5-1: Self Test Failure Codes  
Power  
0
1
2
3
OK  
5 V Failure  
15 V Failures  
Both Failed  
Analog  
0
1
2
3
OK  
DAC A (0–1 V Concentration)  
DAC B (0–1 V Range ID)  
Both Failed  
Teledyne Analytical Instruments  
5-2: Part I  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Part I: Control Unit  
Maintenance 5  
Preamp  
0
1
2
3
OK  
Zero too high  
Amplifier output doesn't match test input  
Both Failed  
5.3 Major Internal Components  
The major components in the Control Unit are shown in Figure 5-3.  
Outer Door  
Inner Door  
Main PCB  
PreampPCB  
Display PCB  
Doors shown removed  
for clarity  
Electrical Connector Pane  
Box Subassembl  
(Gas Connector  
Panel not shown.  
Figure 5-3: Control Unit Major Internal Components  
Part I: 5-3  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5 Maintenance  
Model 3010MB Percent Paramagnetic Oxygen Analyzer  
WARNING: HAZARDOUS VOLTAGES EXIST ON CERTA1IN  
COMPONENTS INTERNALLY WHICH MAY PERSIST  
FOR A TIME EVEN AFTER THE POWER IS TURNED  
OFF AND DISCONNECTED.  
The 3010MB Control Units contain the following major components:  
Power Supply  
Motherboard (with Microprocessor, RS-232 chip, and  
PreamplifierPCB)  
Front Panel Display Board and Displays—  
5 digit LED meter  
2 line, 20 character, alphanumeric, VFD display  
See the drawings in the Drawings section in back of this manual  
fordetails.  
Teledyne Analytical Instruments  
5-4: Part I  
Download from Www.Somanuals.com. All Manuals Search And Download.  
OPERATING INSTRUCTIONS  
Model 3010M  
Oxygen Analyzer  
Part II: Analysis Unit  
Part Number D-71394  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 3010M Oxygen Analyzer  
Table of Contents  
1 OperationalTheory  
1.1 Introduction .................................................................... 1-1  
1.2 Precise Paramagnetic Sensors...................................... 1-1  
1.2.1 Principles of Operation .......................................... 1-1  
1.3 Cross Intrface ................................................................. 1-3  
2 Maintenance  
2.1 Routine Maintenance..................................................... 2-1  
2.2 Major Internal Components............................................ 2-1  
2.3 Sensor Replacement ..................................................... 2-1  
2.4 Output Goes Negative.................................................... 2-1  
ii: Part II  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Paramagnetic Oxygen Analyzer  
Operational Theory 1  
Operational Theory  
1.1 Introduction  
The analyzer is composed of three subsystems:  
1. ParamagneticSensor  
2. SampleSystem  
3. Electronic Signal Processing, Display and Control  
The sample system is designed to accept the sample gas and transport it  
through the analyzer without contaminating or altering the sample prior to  
analysis. The Paramagnetic Sensor is an electromechanical device that  
translates the amount of oxygen present in the sample into an electrical  
signal. The electronic signal processing, display and control subsystem  
simplifies operation of the analyzer and accurately processes the sampled  
data. The microprocessor controls all signal processing, input/output and  
display functions for the analyzer.  
1.2 Precise Paramagnetic Sensor  
1.2.1 Principles of Operation  
The heart of the 3010M is a paramagnetic type oxygen sensor that is  
maintanance free and has a long lifetime. Oxygen has a very high magnetic  
sucseptibility compared to other gases and thus displays a particularly para-  
magnetic behavior. A small glass dumbbell filled with nitrogen and rotating  
on a taut platinum wire is suspended in an inhomogneous magnetic field.  
This glass dumbbell is diamagnetic and tends to rotate out of the magnetic  
field. The strength of the resulting torque determined by an susceptibility of  
the sample gas. This torque is compensated for by a counter torque induced  
by an electrically charged platinum coil on the dumbbell. The zero position  
of the dumbbell is controlled by means of an optical system consisting of a  
light source, a mirror at the dumbbell axis and a pair of detectors. The  
difference between the compensating currents required to bring the dumb-  
TeledyneAnalyticalInstruments  
Part II: 1-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1 Operational Theory  
Model 3010M  
bell to the zero position in the presence of zero gas (i.e. no O2 present) or of  
sample gas is proportional to the partial pressure of oxygen in the sample  
gas.  
The sensor enclosure is temperature controlled to 55 degrees centigrade  
to insure that the magnetic susceptibility of oxygen in the sample is not  
affected by the ambient temperature. The measuring cell has a measuring  
volume of approximately 3 cm3  
Sensor  
(side view shown)  
The Paramagnetic sensor enclosure holds not only the sensing elements,  
butthetemperaturecontrollerelectronics, heatingelements, sensorelectron-  
ics, and preamplifier. The Electronics and heating elements require a separate  
power source, from the rest of the 3020M capable of delivering 1.5 amps  
approximately at 24 volts dc. The output of the sensor is roughly calibrated  
to be 0 to 1 volt DC for the the range of 0 to 100 % 02. The true calibration  
of the sensor is carried out by the microprocessor as described later in chap-  
ter 4. The electrical interconnections to the sensor are done through a 15 pin  
D connector. Some signals from the sensor are not connected. They are only  
useful for troubleshooting, by trained personnel, as test points.  
TeledyneAnalyticalInstruments  
1-2: Part II  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Paramagnetic Oxygen Analyzer  
Operational Theory 1  
Sensor  
(rear view shown)  
Pin out:  
1
2
3
4
5
6
7
8
-15Vdc test pin (Not connected)  
+15Vdc test pin (Not connected)  
Measuring ground  
Not used  
Preamplifier output ~0- 1 Vdc  
Not used  
24 Vdc return, power ground  
Collector of transistor switching heating element (Not con-  
nected)  
9
Fault signal: <0 Vdc= OK, >4.6 Vdc = Fault (Not connected)  
Not used  
10  
11  
12  
13  
14  
15  
Negative reference voltage (Not connected)  
Positive reference voltage (Not connected)  
Nominal temperature voltage signal (Not connected)  
Actual temperature voltage signal (Not connected)  
+24 Vdc power.  
1.3  
Cross Interference  
As mentioned previously, the selectivity of the measuring system is  
based on the extraordinarily high magnetic susceptibility of Oxygen com-  
pared to other gases. In most cases the presence of other gases can be ig-  
nored but significant errors can occur when the sensor is calibrated with a  
mixture of oxygen and nitrogen and the sample gas consists mainly of other  
gases with considerable magnetic susceptibility. In this case, the reading  
shows a measured value even if the sample gas contains no Oxygen. It is  
actually displaying the cross sensitivity to another gas.  
The following table shows the cross sensitivity of some gases when  
changing from pure nitrogen to 100% of one of the gases listed.  
TeledyneAnalyticalInstruments  
Part II: 1-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1 Operational Theory  
Model 3010M  
Gas  
Cross  
Gas  
Cross  
Sensitivity  
in vol. %  
Sensitivity  
in vol. %  
AcetyleneC2H2  
AlleneC3H4  
AmmoniaNH3  
ArgonAr  
BromineBr2  
1.2ButadieneC4H6  
1.3ButadieneC4H6  
n-ButaneC4H10  
I-ButaneC4H8  
cis 2-Butane C4H8  
trans2-ButaneC4H8  
CarbondioxideCO2  
CarbonmonoxideCO  
ChlorineCl2  
-0.24  
-0.44  
-0.26  
-0.22  
-1.30  
-0.49  
-0.49  
-1.11  
-085  
-0.89  
-0.92  
-0.27  
+0.06  
-0.77  
-1.56  
-0.43  
-0.26  
+0.30  
-2.10  
-1.70  
+0.24  
-.0.61  
HydrogenchlorideHCI -0.30  
HydrogenfluorideHF  
HydrogeniodideHI  
+0.10  
-1.10  
HydrogensulphideH2S -0.39  
IodineI  
-2.40  
-1.11  
-1.49  
-0.51  
-0.20  
-0.20  
+0.13  
-1.49  
+0.43  
+28.00  
+40.00  
-2.50  
-1.45  
-0.86  
-0.55  
-0.63  
-0.02  
-0.95  
IsobutaneC4H10  
IsopantaneC5H12  
Krypton Kr  
LaughinggasN2O  
MethaneCH4  
NeonNe  
NeopentaneC5H12  
Nitric acid HNO3  
NitrogendioxideNO2  
Nitrous oxide NO  
n-OctaneC8H18  
n-PentaneC5H12  
PropaneC3H8  
PropyleneC3H6  
Vinyl chloride  
WaterH2O  
CyclohexaneC6H12  
EthaneC2H6  
EthyleneC2H4  
HeliumHe  
n-HeptaneC7H16  
n-HexaneC6H14  
HydrogenH2  
HydrogenbromideHBr  
XenonXe  
With gas mixtures the components are weighted according to their  
proporationalvolumes.  
The cross interference with a gas mixture can be determined in advance if the  
concentration of the individual background gases is known.  
Example:  
Cross sensitivity calculation for a gas mixture:  
The gas to be measured consists of 10 % CO2, 40 % of N2, and 50%  
of Ar.  
C02: -0.27 x 10% = -0.027  
Ar: -0.22x5O%= -0.11  
TOTAL CROSS INTERFERENCE -0.137  
Calculation of the zero value:  
(0% 02) - (Cross Interference Value) = Zero Point Value  
0%-(-0.137) = +0.137%02  
TeledyneAnalyticalInstruments  
1-4: Part II  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Paramagnetic Oxygen Analyzer  
Maintenance 2  
Maintenance  
2.1 Routine Maintenance  
Aside from normal cleaning and checking for leaks at the gas connec-  
tions, routine maintenance is limited to replacing Paramagnetic Sensor and  
fuses, and recalibration. For recalibration, see Section 4.4 The Zero and  
Span Functions.  
WARNING: SEE WARNINGS ON TITLE PAGE OF THIS MANUAL.  
2.2 Major Internal Components  
The 3010M contains the following major internal components:  
ParamagneticSensor  
2.3 Sensor Replacement  
The Paramagnetic Sensor is maintenance free, and has a long life.  
Should the sensor fail, contact the factory for service and replacement.  
2.4 Output Goes Negative  
If during operation of the instrument, gas pressure exceeds 20 psig, the  
output of the sensor may reverse polarity, readings on the display may go  
negative. If this happens:  
Make sure analyzer is mounted in the upright position (see  
chapter 3, Installation).  
Turn power off by unplugging the instrument.  
Turn gas pressure to less than 20 psig.  
Turn power on again.  
Return instrument to analyze mode.  
Part II: 2-1  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2 Maintenance  
Model 3010M  
2-2: Part II  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Paramagnetic Oxygen Analyzer  
Appendix  
Appendix  
A-1 Specifications  
Packaging: Bulkhead mount.  
Sensor: ParamagneticSensor.  
Ranges: Three user definable ranges from 0-5 % to 0-  
100 %.  
Autoranging with range ID output.  
SampleSystem: Positivepressureservice.  
Vacuumservice(optional).  
Auto Cal / Auto Zero. (Available with op-  
tional,electricallyoperatedvalves.)  
Alarms: One system-failure alarm contact to detect  
powerfailure.  
Twoadjustableconcentrationthresholdalarms  
withfullyprogrammablesetpoints.  
Diagnostics: Start-uporon-demandself-testingfunction.  
Response Time: 90% in 10 seconds or less.  
Displays: 2 line by 20 alphanumeric, VFD screen, and  
one 5 digit LED display.  
DigitalInterface: Full duplex RS-232 communications port.  
Power: Universal power supply 100-240 V ac, at 50  
or 60 Hz, 70 Watts max.  
OperatingTemperature: 0-45 °C  
Accuracy: ±1% of full scale at constant temperature.  
±5% of full scale over operating temperature  
A-1  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix  
Model 3010M  
range, on factory default analysis ranges, once  
thermal equilibrium has been achieved.  
Analog outputs: 0-1 V dcpercent-of-range  
0-1 V dc range ID.  
4-20 mADCpercent-of-range  
4-20 mADC range ID.  
Password Access: Can be user-configured for password  
protection.  
A-2  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Paramagnetic Oxygen Analyzer  
Appendix  
A-2 Recommended 2-Year Spare Parts List  
QTY.  
PART NUMBER  
DESCRIPTION  
1
1
1
1
4
C62371B  
D70046  
C70043  
C62365-A  
F1295  
Display PCB  
CustomerInterfacePCB  
PercentPreamplifierBoard  
Main PCB (4-20 mA)  
Fuse, 4A, 250V, 5x20 mm, T (Slow Blow)  
A minimum charge is applicable to spare parts orders.  
Note: Orders for replacement parts should include the part number (if  
available) and the model and serial number of the instrument for  
which the parts are intended.  
Orders should be sent to:  
TeledyneAnalyticalInstruments  
16830 Chestnut Street  
City of Industry, CA 91749-1580  
Phone (626) 934-1500, Fax (626) 961-2538  
TWX (910) 584-1887 TDYANYL COID  
Web:  
or your local representative.  
A-3 Drawing List  
D-66826 WiringDiagram  
C-71466 InterconnectionDiagram  
A-3  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix  
Model 3010M  
NOTE: The MSDS on this material is available upon request  
through the Teledyne Environmental, Health and  
Safety Coordinator. Contact at (626) 934-1592  
A-4  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Percent Paramagnetic Oxygen Analyzer  
A-5 Zero Cal  
Appendix  
The ZERO function on the MAIN MENU is used to enter the zero  
calibration function. Zero calibration can be performed in either the auto-  
matic or manual mode. In the automatic mode, an internal algorithm com-  
pares consecutive readings from the sensor to determine when the output is  
within the acceptable range for zero. In the manual mode, the operator  
determines when the reading is within the acceptable range for zero.  
Make sure the zero gas is connected to the instrument.  
If you have trouble zeroing, you may need to skip to section 4.8.1.3  
CellFailure.  
AutoModeZeroing  
Select ZERO to enter the ZERO function. The ZERO screen allows  
you to select whether the zero calibration is to be performed automatically or  
manually. Use the DOWN/UP control to toggle between AUTO and MAN  
zero settling. Stop when AUTO appears on the display.  
Zero:Settling:AUTO  
<ENT>ToBegin  
ENTER to begin zeroing.  
#### % Zero  
Slope=#### ppm/s  
The beginning zero level is shown in the upper left corner of the dis-  
play. As the zero reading settles, the screen displays and updates information  
on Slope (unless the Slope starts within the acceptable zero range and does  
not need to settle further).  
Then, and whenever Slope is less than 0.08 for at least 3 minutes,  
instead of Slope you will see a countdown, such as 1 Left, 0 Left. These are  
steps in the zeroing process that the system must complete, AFTER settling,  
before it can go back to ANALYZE.  
#### % Zero  
1Left=### ppm/s  
The zeroing process will automatically conclude when the output is  
within the acceptable range for a good zero. Then the analyzer automatically  
returnstotheANALYZEscreen.  
A-5  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix  
Model 3010M  
ManualModeZeroing  
ENTER the ZERO function. The screen that appears allows you to  
select between automatic or manual zero calibration. Use DOWN/UP to  
toggle between AUTO and MAN zero settling. Stop when MAN appears on  
thedisplay.  
Zero:Settling:Man  
<ENT>ToBegin  
ENTER to begin the zero calibration. After a few seconds the first of  
five zeroing screens appears. The number in the upper left hand corner is the  
first-stage zero offset. The microprocessor samples the output at a predeter-  
mined rate. It calculates the differences between successive samplings and  
displays the rate of change as Slope= a value in parts per million per second  
(ppm/s).  
#### % Zero  
Slope=#### ppm/s  
NOTE:It takes several seconds for the true Slope value to display.  
Wait about 10 seconds. Then, wait until Slope is sufficiently  
close to zero before using ENTER to finish zeroing.  
Generally, you have a good zero when Slope is less than 0.05 ppm/s  
for about 30 seconds. When Slope is close enough to zero, ENTER it. In a  
few seconds, the screen will update.  
Once zero settling completes, the information is stored in the  
microprocessor,andtheinstrumentautomaticallyreturnstotheANALYZE  
screen.  
A-6  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Technicolor Thomson CD Player TM9255 User Manual
Tiger Games 65 135 User Manual
Toshiba Computer Accessories GN 2010 User Manual
Toshiba Tablet PDA0EU00101Y User Manual
Triarch Indoor Furnishings 29274 User Manual
Triarch Indoor Furnishings 29644 User Manual
Tripp Lite TV Cables N001 025 BK R User Manual
Uniden Automobile Accessories AX52 User Manual
Victor Corp Flat Panel Television LT 26S60BU User Manual
Vulcan Hart Fryer 1GR45C ML 136414 User Manual